-------
Figure 6-13. Stage 3 Contaminant Injection Results for Turbidity
aldicarb injections produced changes of 0.82, 0.12, 0.43, and 1.03 ntu, respectively. Because of
the inherent turbidity of an E. coli culture, it was expected that turbidity would be consistently
responsive to that contaminant. Also, because aldicarb was completely dissolved, it was not
expected to increase the turbidity of the water upon injection. However, the conditions
surrounding the injection of both contaminants, such as the potential co-injection of air bubbles,
may have affected the turbidity as much as or more than the contaminant itself. Regardless of
what caused the variations in turbidity, the continuous monitor tracked the relative magnitude of
the change in turbidity rather well for the first E. coli injection and the final aldicarb injection
(data only available for one monitor); but, for the other two injections, the uncertainty of the
background made it difficult to determine if a change occurred.
Figures 6-14 through 6-16 show the effect of the injections on pH, total alkalinity, and
conductivity. The pH was clearly affected by both E. coli injections (as shown by both the
reference and continuous measurements) and, to a lesser extent, by the aldicarb injections. This
was also the case for total alkalinity. Both E. coli injections produced rather sharp downward
spikes in pH and upward spikes in total alkalinity and conductivity. Note that a second E. coli
injection was performed by T&E facility staff (not a part of the ETV test) immediately after the
second E. coli injection. Its effect is observed in all three of these measurements. There was a
downward trend in pH and total alkalinity (for both the reference and TitraSip™ measurement)
during the aldicarb injection. Aldicarb had not altered the pH or total alkalinity during the
Stage 2 injections, so this result was unexpected.
33
-------
8.7
8.6
8.5
8.4
8-3
8.2
8.1
8
7.9
E. coin
E coll 2
Aldicart 3
Aldicart) 4
Figure 6-14. Stage 3 Contaminant Injection Results for pH
Figure 6-15. Stage 3 Contaminant Injection Results for Total
Alkalinity
34
-------
390
E coll1
350
Eco*'2 Aldicarb 3
Aldicarti 4
Each section of this figure represents the time between contaminant injection
and the sensor's return to a baseline measurement (approximately 24 hours)
Figure 6-16. Stage 3 Contaminant Injection Results for Conductivity
6.5 Inter-unit Reproducibility
Two TitraSip™ units were compared throughout the verification test to determine whether they
generated results that were similar to one another. This was done using the TitraSip™ data
collected whenever a reference sample was collected throughout the verification test. Two
evaluations were performed to make this comparison. First, the results from Unit 2 were graphed
on the y-axis, those from Unit 1 were graphed on the x-axis, and a line was fitted to the data.
Second, a t-test assuming equal variances was performed on those same data. For the linear
regression analysis, if both TitraSip™ units reported the identical result, the slope of such a
regression would be unity (1), the intercept zero (0), and the coefficient of determination (r2) 1.0.
The slope can indicate whether the results are biased in one direction or the other, while the
coefficient of determination provides a measure of the variability of the results. The t-test shows
whether the sensors generated statistically similar data. Small p-values (<0.05 at a 5%
confidence level) would suggest that the results from the two units are significantly different
from one another. Table 6-7 gives the slope, intercept, and coefficient of determination for the
inter-unit reproducibility evaluation and the p-value for the t-test performed for each sensor.
As shown in Table 6-7, the pH, temperature, and total chlorine results had coefficients of
determination greater than 0.94 and slopes within 6% of unity, indicating that their results were
very similar and repeatable. Confirming that evaluation, the t-test p-values for the same three
parameters were 0.85, 0.92, and 0.48, respectively, indicating that each unit was generating
statistically similar results. The conductivity sensors had a coefficient of determination greater
than 0.89, indicating that these data were highly correlated with one another; however, the slope
values were approximately 9% and 16% from unity. This reflected the tendency of Unit 1 to
35
-------
Table 6-7. Inter-unit Reproducibility Evaluation
Parameter
Conductivity
pH
Total alkalinity
Turbidity
Temperature
Total chlorine
Slope
1.16
0.94
0.79
0.67
1.06
1.06
Intercept
-38.1
0.545
18.1
0.104
-1.22
0.03
r2
0.896
0.981
0.873
0.683
0.942
0.958
t-test p- value
0.110
0.851
0.149
0.449
0.915
0.481
experience negative spikes throughout the verification test. However, the difference between the
two units was not consistent enough to cause the t-test to indicate a significant difference
between the two units.
The total alkalinity and turbidity results from both units were not statistically different from one
another. Therefore, none of the water quality parameter measurements of the TitraSip™ units
were significantly different. These results were confirmed through visual observation of the
figures throughout Chapter 6 because, when graphed, the data from both units were usually
almost superimposed on each other. The only exception was total alkalinity, for which Unit 2
measurements were consistently higher than Unit 1; but, because of the variability of the Unit 1
measurements, they were not significantly different from each other.
6.6 Ease of Use and Data Acquisition
The TitraSip™ units required daily calibration by the verification staff. Once each day, the
controller was manually rebooted, and an automatic "prime and purge" routine was performed to
prepare the units for the upcoming day's analyses. Analysis routines for the following 24 hours
were programmed at this time. Other daily tasks included checking the levels of total chlorine
and total alkalinity titration reagents and calibration standards for the pH, conductivity, and
turbidity meters. The pH and conductivity meters were calibrated daily by using another
automatic routine, while the turbidity meter was only calibrated once per week. Because the
TitraSip™ units collect a sample from the flowing stream and then perform analyses on that
water sample (conductivity, temperature, pH, turbidity, total alkalinity simultaneously, followed
a few minutes later by total chlorine), they generate a complete set of results approximately
every 30 minutes. Therefore, that is the maximum data collection frequency. These results were
stored in a database that was downloaded into a delimited text file for import into Microsoft®
Excel. The software used to program the calibration and analysis routines was easy to use. Note
that this equipment appeared to be a bench-top instrument (as opposed to a field-deployable
instrument that attaches to a wall).
A month-long period during Stage 3 required in-depth troubleshooting of Unit 1. Initially, the
sample cell on that unit would not drain completely between analyses of separate pH calibration
solutions. The most obvious way in which the problem made itself known was through
decreased pH results (see Figure 6-3) and occasional very high outlying total chlorine results.
Failed conductivity calibrations also made the problem evident. Because the TitraSip™ units
each use a single sample cell for every analysis, any residual sample from a previous calibration
or rinse solution can skew the results of subsequent analyses. The ETV staff worked with the
36
-------
Man-Tech staff to resolve the drain problems. Once resolved, both TitraSip™ units functioned
properly through the end of the verification test. Note that the method for conductivity
calibration was altered near the beginning of Stage 2 of the verification test. This greatly
decreased the percent differences of the conductivity results for the rest of the test.
37
-------
Chapter 7
Performance Summary
Evaluation Parameter
Stage 1 —
Accuracy
Stage 2—
Response to
Injected
Contaminants
Stage 3 —
Accuracy During
Extended
Deployment
Stage 3—
Accuracy After
Extended
Deployment
Stage 3 —
Response to
Injected
Contaminants
Injection
Summary
Inter-unit
Reproducibility
(Unit 2 vs. Unit 1)
Units 1 and 2, range
of %D (median)
Nicotine
Arsenic
trioxide
Aldicarb
Reference
TitraSip™
Reference
TitraSip™
Reference
TitraSip™
Units 1 and 2, range
of %D (median)
Unit 1, %D
Unit 2, %D
E. coli
Aldicarb
Reference
TitraSip™
Reference
TitraSip™
Total
Chlorine
-13.2 to
20.6 (7.5)
-
-
-
-
-
-
-18.0 to
30.0 (2.7)
1.0
0.0
-
-
-
-
Tem-
perature
-9.1 to
52.5 (-0.04)
NC
NC
NC
NC
NC
NC
-15.7 to
3.7 (-3.1)
-2.2
-1.9
NC
NC
NC
NC
Conductivity
37.9 to
94.3 (57.5)(a)
NC
NC
+
+(0
NC
NC
-2.8 to
5.2 (0.7)
0.3
1.1
+
+
NC
NC
pH
-2.2 to
5.4 (0.6)
NC
NC
+
+
NC
NC
-4.4 to
0.7 (-1.1)
-1.0
-2.1
-
-
-
-
Total
Alkalinity
3.2 to
30.4(11.5)
NC
NC
+
+
NC
NC
-16.5 to
14.4 (5.7)
-0.4
4.5
+
+
-
-
Turbidity
-65.2 to
0.6 (-45.2)
(b)
(b)
(b)
(b)
(b)
(b)
-96.7 to 155.3
(-37.3)
35.3
41.2
+
(c)
+
(c)
For a reason that is not clear, aldicarb and total alkalinity altered the pH, as measured by the reference
method, during the Stage 3 injections, but not during the Stage 2 injections.
Slope (intercept)
r2
p-value
1.06 (0.03)
0.958
0.481
1.06 (-1.22)
0.942
0.915
1.16 (-38.1)
0.896
0.110
0.94 (0.545)
0.981
0.851
0.79(18.1)
0.873
0.149
0.67(0.104)
0.683
0.449
All sensors generated results that were similar and repeatable between the units.
Ease of Use and
Data Acquisition
The TitraSip™ units required daily calibration, which involved operator intervention. Initially, the sample
cell on Unit 1 did not drain completely between pH calibration solutions, but once the drain problem was
resolved, both units functioned properly. Monitor results were recorded once every 30 minutes, which is
the maximum data collection frequency.
Calibration procedure for the conductivity meter was changed after Stage 1, resulting in much lower percent
differences throughout the remainder of the verification test.
Relatively large uncertainties in the reference and continuous measurements made it difficult to determine a
significant change.
Duplicate injection results did not agree.
+/- = Parameter measurement increased/decreased upon injection.
NC = No obvious change was noted through a visual inspection of the data.
(b)
(c)
38
-------
Chapter 8
References
1. U.S. EPA, EPA Method 150.1, pH, in Methods for Chemical Analysis of Water and Wastes,
EPA/600/4-79-020, March 1983.
2. American Public Health Association, et al., SM 2510, Conductivity, in Standard Methods for
the Examination of Water and Wastewater, April 13, 2004.
3. American Public Health Association, et al., SM 2320B, Alkalinity by titration, in Standard
Methods for the Examination of Water and Wastewater, April 13, 2004.
4. American Public Health Association, et al., SM 4500-C1 B, Chloride by silver nitrate
titration, in Standard Methods for the Examination of Water and Wastewater, April 13, 2004.
5. U.S. EPA, EPA Method 170.1, Temperature, in Methods for Chemical Analysis of Water and
Wastes, EPA/600/4-79-020, March 1983.
6. American Public Health Association, et al., SM 2130B, Turbidity, nephelometric, in
Standard Methods for the Examination of Water and Wastewater, April 13, 2004.
7. Test/QA Plan for Verification of Multi-Parameter Water Monitors for Distribution Systems,
Battelle, Columbus, Ohio, August 2004.
8. U.S. EPA, EPA Method 310.1B, Alkalinity—Titrimetric, pH 4.5, in Methods for Chemical
Analysis of Water and Wastes, EPA/600/4-79-020, March 1983.
9. American Public Health Association, et al., SM 4500-G, Residual Chlorine, in Standard
Methods for the Examination of Water and Wastewater, 19th Edition, Washington, D.C.,
1997.
10. Quality Management Plan (QMP)for the ETV Advanced Monitoring Systems Center,
Version 5.0, U.S. EPA Environmental Technology Verification Program, Battelle,
Columbus, Ohio, March 2004.
39
-------