Version 0, August 30, 2006 Environmental Technology Verification Generic Verification Protocol for Biological and Aerosol Testing of General Ventilation Air Cleaners EPA Cooperative Agreement R-83191101 Research Triangle Institute Project 09309 Prepared by: Air Pollution Control Technology Center Research Triangle Institute Research Triangle Park, NC Approved by: APCT Center Director: Signed by Andrew Trenholm, August 21, 2006 Andrew Trenholm APCT Center Quality Manager: Signed by Gary Eaton, August 21, 2006 W. Gary Eaton EPA Project Officer: Signed by Michael Kosusko, August 30, 2006 Michael Kosusko EPA Quality Manager: Signed by Paul Groff, August 30. 2006 Paul Groff KRTI INTERNATIONAL ------- Research Triangle Institute Page ii Generic Verification Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners This page left intentionally blank. ------- Research Triangle Institute Page iii Generic Verification Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners TABLE OF CONTENTS Table of Contents iii List of Figures v List of Tables v List of Acronyms/Abbreviations/Definitions vii 1.0 Introduction 1 1.1: Environmental Technology Verification Program 1 1.2: Air Pollution Control Technology Verification Center 2 2.0: Verification Description 2 2.1: Identification and Acquisition of Devices 3 2.2: Performance of ASHRAE 52.2-1999 Test 4 2.3: Performance of Culturable Bioaerosol Testing 4 2.4: Performance of Inert Particle Testing 4 2.5: Preparation of Report 5 2.6: Data Quality Objectives and Criteria for Measurement Data 5 2.7: Special Training Requirements/Certification 6 2.8: Documentation and Records 6 2.8.1: Laboratory Documentation 6 2.8.2: Reporting 7 2.8.3: Verification Reports and Verification Statements 7 3.0 Test Program 8 3.1: Test Design 8 3.1.1: ASHRAE 52.2-1999 and Inert Testing 8 3.1.2: Culturable Bioaerosol Testing 10 3.2: Sampling Methods Requirements 12 3.3: Sample Handling and Custody Requirements 12 3.4: Analytical Methods Requirements 12 3.5: Quality Control Requirements 12 3.6: Instrument/Equipment Testing, Inspection, and Maintenance Requirements 12 3.7: Instrument Calibration and Frequency 12 3.8: Inspect!on/Acceptance Requirements for Supplies and Consumables 13 3.9: Data Management 13 3.9.1: Data Recording 13 3.9.2: Data Analysis 13 3.9.3: Data Storage and Retrieval 15 ------- Research Triangle Institute Page iv Generic Verification Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners References 16 Appendix A: Test Specifications 18 Appendix B: Inert Aerosol Run Sheet 24 Appendix C: Bioaerosol Run Sheet 26 LIST OF FIGURES Figure Al. Schematic of test duct (top view) used for device testing 18 LIST OF TABLES Table 1. DQOs for Inert Aerosol Tests 5 Table 2. DQOs for Filtration Efficiency for Culturable Bioaerosol 5 Table Al. Quality Control Parameters for Inert Aerosol Tests 19 Table A2. Quality Control Parameters Associated with Scanning Mobility Particle Sizer (SMPS) and Conditioning Aerosol 21 Table A3. Quality Control Parameters for Bioaerosols 22 Table A4. Quality Control Parameters for the Test Duct 24 ------- Research Triangle Institute Generic Verification Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners List of Acronyms/Abbreviations/Definitions Page v ADQs AGI ANSI ASHRAE ASME cm3 CPU cfm cm CPC culturable CV dso DQO EI, £2, £3 electret EPA Eq. ETV g HEPA in. ISO KC1 Kr L m3 MERV min min. ML m mL mm MSDS MS2 audits of data quality all glass impinger American National Standards Institute American Society of Heating, Refrigerating, and Air-conditioning Engineers American Society of Mechanical Engineers cubic centimeter(s) colony forming unit cubic feet per minute centimeter(s) condensation particle counter able to be grown on microbiological media coefficient of variance 50% cut point on Andersen sampler data quality objective average minimum particle-size efficiency designator of ASHRAE a filter comprised of fibers that contain an embedded electrostatic 52.2 charge U.S. Environmental Protection Agency equation Environmental Technology Verification Program gram(s) high efficiency particulate air inch(es) International Organization for Standardization potassium chloride Krypton liter(s) cubic meter(s) minimum efficiency reporting value of ASHRAE 52.2 minimum minute(s) microbiology laboratory meter(s) milliliter(s) millimeter(s) material safety data sheet bacterial virus or bacteriophage ------- Research Triangle Institute Generic Verification Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page vi OPC PCO PEs PFU PSE psig PSL QA QC QM QMP RH RTI sec SMPS SOP t T/QAP TSAs //m optical particle counter photocatalytic oxidation performance evaluations plaque forming unit particle size (removal) efficiency pounds per square inch gauge polystyrene-latex quality assurance quality control quality manager quality management plan relative humidity Research Triangle Institute second(s) scanning mobility particle counter standard operating procedure temperature test/quality assurance plan technical system audits micrometer ------- Research Triangle Institute Page 1 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners 1.0 INTRODUCTION The U.S. Environmental Protection Agency (EPA) established the Environmental Technology Verification Program (ETV) in 1995. Under a cooperative agreement with EPA/ETV, Research Triangle Institute (RTI) operates the Air Pollution Control Technology Verification Center (APCT), and developed this protocol to verify filtration efficiency and bioaerosol collection and/or inactivation efficiency of heating, ventilation and air conditioning (HVAC) air cleaners for culturable bioaerosol and aerosol challenges. This protocol describes the ETV considerations and requirements for verification of in-duct air cleaners based on filtration, ultraviolet (UV) illumination, photocatalytic oxidation (PCO) and combinations of those technologies. At this time the test plan is for filters, however an addendum will be added as needed to test other devices. The protocol can be used for UV, PCO and other technologies if they are compatible with the test facilities and procedures of the protocol. It is anticipated that the devices tested will be compatible with a nominal 24" x 24" test duct cross section. This protocol is based on previous work for three ETV-related projects. Stakeholder groups were convened under these projects to provide input into the selection of technologies and into the development of protocols and test/quality assurance plans (T/QAPs). • Under the ETV Indoor Air Pilot, a test protocol1 and test plan2 were developed and validated for general ventilation media devices. • As part of the ETV Safe Buildings for homeland security, a test protocol3 and test plan4 developed which included bioaerosol testing. • Recently, the EPA Technology Testing and Evaluation Program (TTEP) developed a test plan for testing UV light systems used in ventilation ducts for bioaerosols5. The methods and procedures in these documents were supplemented based on RTFs experience conducting testing for commercial clients for bioaerosols and testing based on American National Standards Institute (ANSI)/American Society of Heating, Refrigerating and Air- Conditioning Engineers, Inc. (ASHRAE) Standard 52.2-19996, Method of'Testing General Ventilation Air-Cleaning Devices for Removal Efficiency by Particle Size (ASHRAE 52.2-1999). 1.1 Environmental Technology Verification EPA through its Office of Research and Development (EPA-ORD) instituted the ETV Program to verify the performance of innovative and improved technical solutions to problems that threaten human health or the environment. EPA created the ETV Program to accelerate the entrance of new and improved environmental technologies into the marketplace. It is a voluntary, nonregulatory program. Its goal is to verify the environmental performance characteristics of commercially ready technologies through the evaluation of objective and quality-assured data so that potential purchasers and permitters are provided with an independent and credible assessment of what they are buying and permitting. The ETV Program does not conduct technology research or development. ETV test results are always publicly available, and the applicants are strongly encouraged to ensure prior to ------- Research Triangle Institute Page 2 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners beginning an ETV test that they are satisfied with the performance of their technologies. Within the ETV Program, this state of development is characterized as "commercially ready." The provision of high-quality performance data on a commercial technology encourages more rapid implementation of that technology and consequent protection of the environment with better and less expensive approaches. 1.2 Air Pollution Control Technology Verification Center EPA's partner in the Air Pollution Control Technology Verification Center (APCT) is the Research Triangle Institute (RTI), a nonprofit contract research organization with headquarters in Research Triangle Park, NC. The APCT verifies the performance of commercially ready technologies used to control air pollutant emissions. In addition to indoor air quality, APCT include technologies for controlling particulate matter (PM), volatile organic compounds (VOCs), nitrogen oxides (NOx), and hazardous air pollutants (HAPs) from both mobile and stationary sources. The activities of the APCT are conducted with the assistance of stakeholders from various interested parties. Overall, APCT guidance is provided by the Stakeholders Advisory Committee (SAC), whereas the detailed development of individual technology ETV protocols is conducted with input from technical panels focused on each technology area. 2.0 VERIFICATION DESCRIPTION This ETV protocol describes the test and QA procedures that will be used to provide data on the removal efficiency of bioaerosols and inert aerosols by general ventilation air cleaners. While data and methods are available for measuring single-pass inert particle removal efficiencies of air cleaners and filters, no standard method exists for evaluating culturable bioaerosol reduction by these devices. RTI has developed a test method for measuring culturable bioaerosol filtration efficiencies of devices ranging from a room air cleaner to duct-mounted ventilation filters to vacuum cleaner filters7'8'9. Additionally, the method was used in a previous ETV project for Biological Testing of General Ventilation Filters (EPA Contract No. GS10F0283K-BPA-1; Task Order 1101. Research Triangle Institute Project 08787.001)4. The methods discussed in the previous paragraph are the basis for the bioaerosol test and the sub- 0.3 |im inert particle tests in this protocol. Inert particle efficiency tests are used also as a point of comparison for quality assurance (QA)/quality control (QC) of the culturable bioaerosol results and will be used as a "self-consistency" check within the QA framework for the bioaerosol tests. This protocol describes the following tests: 1. A complete ASHRAE 52.2-1999 test, 2. Bioaerosol test with four culturable microorganisms, 3. Inert particle tests, using potassium chloride (KC1) aerosol: a. 0.3 - 10 |im, with optical particle counter (OPC) measurement, b. 0.03 - 0.3 |im, with scanning mobility particle counter (SMPS) measurement, and c. 0.03 - 10 |im, with combined SMPS and OPC measurements. ------- Research Triangle Institute Page 3 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners The above tests will be performed on clean, conditioned and dust-loaded devices as defined in this paragraph. For options 2 and 3, conditioning with the sub-micron aerosol and dust loading with ASHRAE dust will be required for all devices that incorporate filter media. Filter media is the fibrous material used in air filters and other air cleaners to remove particles via filtration. Examples include microglass and polypropylene fibers. Filter media is typically assembled in the form of a flat panel, pleated or bag configuration. Both conditioning and dust loading of filters are done so that the testing better represents what happens when filters are used in real life. Most media filters increase in efficiency as they are used and dust collects on the media. Electret filters have an electrostatic charge applied during manufacturing. These filters may decrease in efficiency at the start of their use cycle and, possibly, throughout the use cycle if the filters are replaced before significant loading (dust accumulation) occurs. For ASFIRAE 52.2-1999, the conditioning step (also called the first dustload) challenges the filter with ASFIRAE dust (a mixture of carbon black, cotton linters, and Arizona road dust) until either a pressure drop increase of 10 Pa is achieved or 30 g of dust is fed, whichever one comes first. Relative to the 25%, 50%, 75% and 100% dust load steps used in the ASHRAE 52.2-1999 test, this is a very low dust challenge intended to simulate dust loading during the early states of the filter's use. The dust loading with ASFIRAE dust is used to simulate the changes that can occur as a media filter accumulates dust; this usually results in an increase in efficiency, but can lead to an efficiency decrease if the filter sheds the dust instead of retaining it. The recommended submicron conditioning step was recently developed as a way to more closely mimic the actual drop-off in efficiency as seen for electret media. Where the ASFIRAE 52.2- 1999 conditioning step may show a decrease in efficiency for these filters, it is usually much less than that shown in situ and is unlikely to show a drop-off for the larger particle sizes even though many media filters do drop-off in efficiency across the entire particle size range. Thus this step is needed to simulate real use efficiencies. Device-specific handling or additional measurements may be necessary for some devices. All of these tests will be performed in a fully qualified ASHRAE 52.2-1999 test duct. This test duct operates at positive pressure to minimize infiltration of room air or bioaerosol. The KC1 aerosol used for the ASFIRAE 52.2-1999 and other inert aerosol tests and the bioaerosols are injected upstream of a mixing baffle to provide aerosol mixing with the airstream. Bioaerosol and inert aerosol concentrations are measured both upstream and downstream of the test section where the air cleaner is installed to obtain the challenge and penetrating concentrations, respectively. The bioaerosol test includes inert testing, as applicable, for QA/QC and reporting purposes. 2.1: Identification and Acquisition of Devices Devices will be selected by the manufacturers and shipped to the testing facility. The full name and description of the product will be provided. If a media device, a separate filter (or device) will be provided for each ASFIRAE 52.2-1999 test; the bioaerosol tests and inert test may be performed on the same filter if desired. For devices with filters, each manufacturer will provide ------- Research Triangle Institute Page 4 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners a backup filter to be used if, for example, the other filters are damaged in transit. A custodian of the devices will be responsible for storage, labeling, etc. of the devices. For non-media devices, which tests are appropriate and how many units are needed will need to be determined on an individual basis in consultation between the manufacturer and the testing personnel. 2.2: Performance of ASHRAE 52.2-1999 Test The ASHRAE 52.2-1999 test will be performed per the standard and will establish the minimum efficiency reporting value (MERV) and other parameters as required in the standard. ASHRAE designed the MERV to represent a filter's minimum performance over multiple particle sizes. In general, a higher MERV indicates higher filter efficiency. Most commercial filters and high end residential filters are now marketed using the MERV. The filtration efficiencies (average of the minimum composite efficiency) are presented by particle size groupings: El, 0.3 to 1.0 //m; E2, 1.0 to 3.0 //m; and E3, 3.0 //m to 10 //m. If other tests are performed on the same model of device, performing the ASHRAE 52.2-1999 test should yield a consistent set of MERV, inert efficiency and bioaerosol efficiency measurements for a device type. 2.3 Performance of Culturable Bioaerosol Testing Biological testing will be performed using four different bioaerosols and one inert aerosol, if appropriate. If ASHRAE 52.2-1999 testing is also performed, a second device will be used for the bioaerosol testing if the device contains media. First, the initial efficiency will be determined using both the biological and the inert aerosols. The inert testing will cover the typical ASHRAE 52.2-1999 particle size range of 0.3-10 jim. If the device contains media, the initial efficiencies will be followed by • Submicrometer conditioning (if applicable, see Section 2.0) and the biological efficiency will be determined after conditioning, • The device will then be loaded with ASHRAE test dust (if applicable, see Section 2.0) to obtain the final pressure drop as appropriate based on the MERV estimated by the inert particles or as chosen by the manufacturer. The bioaerosol efficiencies after dust-loading will be determined. Table 12-1 of the ASHRAE 52.2-1999 standard provides information on the minimum final resistance to be used for each MERV value. The specifics of the testing will be discussed in further detail in Section 3.1.2. 2.4 Performance of Inert Particle Testing Inert testing will be performed using KC1 aerosol on the device when (1) clean and, if applicable (see Section 2.0), (2) conditioned and (3) fully dust-loaded. The device is fully dust-loaded when the minimum final resistance is reached as specified in Table 12-1 of the standard. If ASHRAE 52.2-1999 testing is performed, a different device will be used for this testing if the device contains media. If the bioaerosol testing is performed, this testing may be interspersed with that testing. First, the initial efficiency will be determined using KC1 generated by the ASHRAE 52.2 method and by a Collison nebulizer. The standard generation method will be used with testing with the OPC covering 0.3-10 jim; the nebulizer will be used with the SMPS to cover the particle size range of 0.03-0.3 i^m. If applicable (see Section 2.0), the initial efficiencies will be followed by: • Submicrometer conditioning with the inert efficiency to be determined after conditioning and between conditioning steps, ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners PageS • The device will be loaded with ASHRAE test dust to obtain the final pressure drop as appropriate based on the MERV estimated by the inert particles (Table 12-1 of the standard) or as chosen by the manufacturer; the inert efficiencies after dust-loading will be determined. The specifics of the testing will be discussed in further detail in Section Bl.l. 2.5: Preparation of Report The final step is to complete the verification report and verification statement for each product and test performed and submit them to the EPA. 2.6: Data Quality Objectives and Criteria for Measurement Data Data quality objectives (DQOs) are qualitative and quantitative statements designed to ensure that the type, quality, and quantity of data used are appropriate for the intended application. The DQOs for the critical measurements are found in Tables 1 and 2. The test specifications are found in Appendix A. Table 1. DQOs for Inert Aerosol Tests Parameter Frequency and description DQO OPC (optical particle counter): Penetration error limit for OPC data Each test. Statistical check of data quality. Expected to be achieved on tests of clean air cleaners. May not always be achieved with dust-loaded air cleaners if the air cleaner sheds a significant amount of the collected dust. Per definitions and procedures of ASHRAE 52.2 Section 10.6.46 a — -:= < 0.07P or 0.05 whichever is greater for 0.3 — 3 fjm ^ 0.15P or 0.05 -whichever is greater for 3 - 5.5 /urn ^ 0.20P or 0.05 whichever is greater for 5.5 — 10 T = T-distribution variable, n = number of samples, P = penetration (fraction), o = standard deviation Table 2. DQOs for Filtration Efficiency for Culturable Bioaerosol Parameter Minimum upstream counts for samplers Maximum counts for samplers 100% Penetration (correlation test) Upstream CPUs Frequency and description Each efficiency test. Each efficiency test. Performed at least once per test sequence per organism. Each test. Statistical check of data quality. Control Limits Minimum of 10 CFUVplate or PFUb/plate Maximum of 500 CFU/plate or 800 PFUb/plate Test Acceptable Organism Penetration Range B. atrophaeus 0.85 to 1.15 & A,versicolor S. marcescens 0.80 to 1.20 MS2 0.75 to 1.25 cv1: < 0.25 ------- Research Triangle Institute Page 6 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Parameter Upstream PFUs Frequency and description Each test. Statistical check of data quality. Control Limits CVC< 0.35 a CPU = colony forming units b PFU = plaque forming unit °CV = coefficient of variance All data will be reviewed for accuracy (correctness) and reasonableness. If the results are deemed unreasonable by the senior technical staff (e.g., internally inconsistent), they will be discarded, the procedures reviewed, and the test repeated if necessary. Occasional data points within a test are obvious outliers and will be discarded based on the statistical tests described in and/or referenced by ASTM Standard Practice E 178-02, Standard Practice for Dealing with Outlying Observations10 without requiring the entire test to be repeated. While exact agreement is not expected (due to the different measurements devices) similar results are expected. 2.7: Special Training Requirements/Certification There are no specialized certification requirements specified for these tests. The method chosen for analysis of the inert aerosol particle size efficiency of ventilation devices in the laboratory is restricted to use by, or under the supervision of, personnel experienced in the use of an OPC, SMPS and skilled in the interpretation of raw count data. In addition, for the bioaerosol tests, personnel should have completed at least one formal microbiology course (college or professional/society sanctioned) and gone through extensive informal laboratory training in the microbiology techniques needed for this task. 2.8: Documentation and Records This section identifies the documents and reports to be generated as part of the verification program and the information to be included in the verification reports and verification statements. A description of the data management system established for this task is presented in Section 3.9. Requirements for record keeping and data management for the overall program are found in the U.S. EPA, Environmental Technology Verification Program Quality Management Plan 11 2.8.1: Laboratory Documentation ASHRAE 52.2-1999 and Inert Aerosol Tests The test operator for the inert aerosol test will record the test data and run notes on test run sheets prepared specifically for these tests (An example is presented in Appendix B.) The sheets will be kept in a labeled three-ring binder. The run sheets are designed to prompt the test operator for all required test information: • Testing date, time, and operator; • Manufacturer and model number of device; • Physical description of the device; • QA checks on the equipment and data; and • Test conditions (temperature, relative humidity, atmospheric pressure, air flow rate, device pressure drop). ------- Research Triangle Institute Page 7 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners The particle count data generated by the OPC are recorded by the computer. The file will be saved to the hard drive and later copied to a floppy disk or shared directory for backup. Bioaerosol Tests The bioaerosol test operator will record the test data and notes on a bioaerosol test run sheet (presented in Appendix C). The sheets are kept in a labeled three-ring binder. The run sheets are patterned from the inert aerosol run sheets and designed to prompt the test operator for all required test information: • Device and run number; • Testing date and operator; • Test conditions (t, RH, ambient pressure, air flow rate, pressure drop across ASME nozzle; • Biological suspension information (test organism, suspension preparation, drying air, nebulizer pressure, initial volume, and time on); • Biological sampling scheme (time run begins, sample length, and media); and • Rotameter readings showing the flow rate through the bioaerosol sampler. The organism counts are entered in the project notebook or recorded by a computer. If recorded to a computer, the file will be saved to the hard drive and later copied to a floppy disk or shared directory for backup. 2.8.2: Reporting After the completion of verification tests, the control test data, sample inventory logs, calibration records, and certificates of calibration will be stored. Calibration records will include such information as the instrument being calibrated, raw calibration data, calibration equations, analyzer identifications, calibration dates, calibration standards used and their traceabilities, identification of calibration equipment used, and the staff conducting the calibration. Final reports of self-assessments and independent assessments (i.e., technical systems audits, performance evaluations, and audits of data quality — TSAs, PEs, and ADQs — will be retained. Each verification report and verification statement will contain a QA section, which will describe the extent that verification test data comply with DQOs. 2.8.3: Verification Reports and Verification Statements Verification reports and verification statements will be prepared, reviewed and submitted to the EPA for approval. Procedures for the preparation, review, and dissemination of verification reports and verification statements are described in the U.S. EPA, Environmental Technology Verification Program Quality Management Plan11. It is anticipated that the verification reports and statements will include the filtration efficiency and/or bioaerosol collection/inactivation efficiency of the tested device for the challenges used for the clean, and when containing a filter the conditioned and dust-loaded, device. The following information will be included in the verification reports and verification statements, depending on which test(s) were performed: ------- Research Triangle Institute Page 8 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners • The fractional filtration efficiency of the air device over the 0.03 - 10 i^m size range for the device when (1) clean, and if applicable (see Section 2.0) (2) conditioned and (3) fully dust- loaded; • The collection/inactivation efficiency for the four bioaerosols; • The pressure drop across the clean and (if applicable, see Section 2.0), fully dust-loaded device; • The test air flow rate; • The measured MERV and the associated El, E2, and E3 values of the ASHRAE 52.2-1999 test; • A complete ASHRAE 52.2 report, and • A description and photograph of the device tested. 3.0 MEASUREMENT/DATA ACQUISITION 3.1: Test Design Under this protocol, the following tests may be performed: 1. A complete ASHRAE 52.2-1999 test, 2. Bioaerosol test with four culturable microorganisms, 3. Inert particle tests, using KC1 aerosol: a. 0.3 - 10 jam, with OPC measurement, b. 0.03-0.3 |im, with SMPS measurement, and c. 0.03- 10 jam, with combined SMPS and OPC measurements. The above tests will be performed on clean, conditioned and dust-loaded devices as applicable. Conditioning with the sub-micron aerosol and dust loading will be required for all devices that incorporate filter media. All tests will be performed on devices at an air flow rate acceptable under ASHRAE 52.2-1999. 3.2: ASHRAE 52.2-1999 and Inert Testing The ASHRAE 52.2-1999 test will be run in accordance with the ASHRAE 52.2-1999 test method. A second device will be tested with modified ASHRAE 52.2-1999 procedures to extend the measurements to smaller particle sizes and to condition electret media. All the inert aerosol tests will use laboratory-generated KC1 particles dispersed into the airstream as the test aerosol. A particle counter will measure and count the particles in a series of size ranges both upstream and downstream of the test devices for its efficiency determinations. To simulate the effects of dust accumulation on the devices, the devices will be tested when clean and, if applicable (see Section 2.0), when conditioned and when fully dust-loaded. The dust-loading will follow ASHRAE 52.2-1999 procedures as applicable. Particle Counters For the inert aerosol filtration efficiency measurements, the particle sizing measurements will be made with two particle counting instruments: a Climet Model 500 spectrometer (OPC) or equivalent OPC covering the particle diameter size range from 0.3 - 10 //m in 12 particle sizing channels and a TSI SMPS or equivalent to cover the range from 0.03 - 0.3 //m. For the conditioning aerosol, a TSI condensation particle counter (CPC) (model 3022A or similar) or the equivalent will be used to monitor the upstream concentration. The CPC will function to ------- Research Triangle Institute Page 9 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners monitor the concentration of the submicrometer-sized particles used for conditioning; it will not aid in measuring the efficiency of the devices. The OPC uses a laser-light illumination source and has a wide collection angle for the scattered light. The OPC's sampling rate is 7.1 L/min. (0.25 cfm). The OPC is equipped to provide a contact closure at the end of each sample and also provides a 15 sec. delay in particle counting after each sample. The contact closure is used to control the operation of electromechanical valve actuators in the upstream and downstream sample lines. The 15 sec. delay allows time for the new sample to be acquired. The SMPS consists of a TSI Model 3080 electrostatic classifier and a TSI Model 3010 or 3022 CPC. Depending upon the quality of the data from any individual test, the SMPS can sometimes reliably quantify particles even smaller than 0.03 //m, and when this is the case, those smaller sizes are reported. The ability to quantify sizes smaller than 0.03 //m is determined as defined in Table A2. A data control parameter for the SMPS requires that the coefficient of variance (CV) on upstream counts be computed for each efficiency test based on the upstream particle counts and that the CV be less than 0.30 before the data is used. The lower size ranges for the SMPS are included in the verification report only if they meet the data control parameter. Particle sizes above 0.3 jim will be measured and reported when there are particle counts that meet the data specifications; however, the aerosol generation system necessary to meet our data standards often does not achieve sufficient particle counts for the larger particles. Inert Aerosol Generation Three aerosol generators will be used for the tests as applicable. These generators are needed to cover the range of particle sizes needed; one for the 0.03 - 0.3 //m tests, one for the 0.3 - 10 //m tests, and one to generate the submicrometer conditioning aerosol. All of the aerosols will be generated from KC1 in aqueous solution. The concentrations of KC1 will vary as will the generation technique to give particles in the needed size ranges. For the 0.3-10 //m efficiency tests, the KC1 solution will be nebulized using a two-fluid (air and liquid) atomizing nozzle (Spray Systems 1/4 J siphon spray nozzle). The full description of the test duct is in Appendix A. The nozzle is positioned at the top of a 0.30 m (12 in.)-diameter, 1.3 m (51 in.)-tall transparent acrylic spray tower. The tower serves two purposes. It allows the salt droplets to dry by providing an approximately 40 sec. mean residence time, and it allows larger particles to fall out from the aerosol. After generation, the aerosol passes through a TSI Model 3054 aerosol neutralizer (Krypton-85 radioactive source) to neutralize any electrostatic charge on the aerosol (electrostatic charging is an unavoidable consequence of most aerosol-generation methods). The KC1 solution is fed to the atomizing nozzle at 1.2 mL/min. by a pump. Varying the operating air pressure of the generator allows control of the output aerosol concentration. For the 0.03 - 0.3 //m tests, the KC1 solution will be nebulized with a Collison nebulizer or Laskin nozzle generator. Both of these devices generate smaller particles than the spray nozzle. Inert Conditioning Procedure for Devices with Media Under Test Options 2 and 3 For the conditioning required for devices with filter media if tested under options 2 or 3, the conditioning aerosol will be produced using a bank of Laskin generators nebulizing a 0.1% KC1 ------- Research Triangle Institute Page 10 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners aqueous solution (1 g KC1 to 1 L of water). Previous measurements have indicated that the resultant aerosol is < 0.1 //m in mean diameter. Periodically during the conditioning portion of the test, the device's efficiency will be measured (0.3 - 10 //m) to determine if the efficiency has fallen to its minimum condition. Once the efficiency is at or near its minimum, the conditioning will cease. The duration of conditioning and the concentration of the conditioning aerosol will be monitored during the test. 3.1.2: Culturable Bioaerosol Testing The bioaerosol testing methodology is based on many of the same principals as the inert methodology. Bioaerosol testing uses the same test rig as the inert aerosol and gas-phase testing. Bioaerosol is generated from a suspension of the test organism and the sampling is achieved using bioaerosol samplers. The use of microorganisms as the challenge aerosol requires that a number of technical issues be addressed. These include: • Measuring the survivability and culturability of the organisms through the aerosol generation and collection process; • Determining whether the test organisms are being aerosolized as singlets with a narrow size distribution; • Generating the bioaerosol challenge in sufficient concentration to maintain the sampling duration within the sample time limits of the bioaerosol sampler; and • Establishing the generation protocol for the test organisms. Test Organisms For devices with a filter, the size and shape of the organisms selected for testing are important because the organisms are aerosolized and their filtration efficiency determined. These organisms naturally vary in both their sizes and shapes. Therefore, there is the need to select organisms that reflect that natural diversity. For devices that inactivate bioaerosols, the test organisms also cover a range of susceptibilities to the various inactivation mechanisms of the device. The bioaerosol tests will be conducted using four organisms: one fungal spore, one spore- forming bacterium, one vegetative bacterium, and one virus. The fungal spore Aspergillus versicolor, a 2 - 3.5 jim sphere, is frequently reported as a causative agent of hypersensitivity pneumonitis and has been isolated from a number of problem buildings. The spore form of the bacteria Bacillus atrophaeus (formerly B. subtilis var niger) is elliptically shaped with dimensions of 0.7 - 0.8 x 1 - 1.5 |im. The organism is a ubiquitous environmental bacterium found at high levels in soil and highly associated with indoor dust. Staphylococcus epidermidis (0.5 - 1.5 |im sphere) is a common gram-positive organism and will be the representative vegetative bacterium. Human viruses are thought to be spread by airborne or droplet transmission. Because human viruses can be expensive and cumbersome to work with, the bacterial virus (bactedophage) MS2 (0.02 - 0.03 //m), having approximately the same aerosol characteristics as a human virus, will serve as a surrogate for the viruses of similar and larger size and shape. Although the individual virus particles are in the submicrometer size range, the test particle size planned for the virus tests will span a range of sizes (polydispersed bioaerosol). This test is not ------- Research Triangle Institute Page 11 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners designed to study the removal efficiencies for single individual virus particles; rather, it is designed to determine the removal efficiencies for virus particles as they are commonly found indoors. A representative challenge would be a polydispersed aerosol containing the phage because: • The aerosols created from sneezing and coughing vary in size from < 1 to 20 //m12, but the largest particles settle out and only the smaller sizes remain in the air for extended periods for potential removal by an air cleaner; • For some viruses (i.e., Coxsackie virus), few viruses have been found associated with the smallest particles13; and • Nearly all 1 - 2 //m particles are deposited in the respiratory tract, while larger particles may not be respired. Bioaerosol Preparation and Generation Bacteria suspension preparation for the aerosolization process requires that the specific test organism be grown in the laboratory and the suspension prepared for aerosol generation in the test rig. The microbial challenge suspensions are prepared by inoculating the test organism on solid or liquid media, incubating the culture until mature, wiping organisms from the surface of the pure culture (if solid media), and eluting them into sterile nebulizing fluid to a known concentration, corresponding to a concentration of approximately 1 x 107 CPUs (colony forming units)/mL. Usually, trypticase soy agar (broth) will be used for the bacteria. Sabourauds Dextrose Agar will be used for the fungus. The phage challenge will be prepared by inoculating a logarithmic phase broth culture of the host bacteria with phage and allowing it to multiply overnight or until the majority of the host bacteria are lysed. The mixture is processed to collect and concentrate the phage. Then, the phage stock is filter sterilized (0.2//m) to remove the bacteria. The phage stock will be used as the challenge aerosol. The concentration of the phage stock will be approximately 1 x 1012 or higher plaque forming units (PFU)/mL. The challenge organism suspensions will be aerosolized using a Collison nebulizer (BGI, Waltham, MA) at 15 psi air pressure. The Collison nebulizer generates droplets with an approximate volume mean diameter of 2 //m. The particle diameter after the water evaporates depends on the solids content of the suspension. Particle size is determined by the size of the suspended particles (if singlets). Upstream and downstream sampling of the bacteria and fungus will be accomplished using one- stage Andersen viable bioaerosol samplers or all glass impingers (AGIs). The phage will be collected in all glass impingers (AGIs). The one-stage Andersen sampler is a 400-hole multiple-jet impactor operating at 28 L/min. The dso (50% cut point on Andersen sampler) is 0.65 //m. After sampling, the petri dishes will be removed from the sampler and incubated at appropriate times and temperatures for the test organism being used. CPUs are then enumerated and their identity confirmed. The AGI is a high velocity liquid impinger operating at a flow rate of 12.3 - 12.6 L/min. The dso is approximately 0.3 //m. The AGI is the sampler against which the other commonly used bioaerosol samplers are often compared. The AGI (containing collection fluid) is plated and the CPUs or PFUs are enumerated. ------- Research Triangle Institute Page 12 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners The experimental conditions and sampling times will be adjusted so that these samplers will be used within their upper and lower sampling limits. To quantify the microbial counts, the plates are incubated at the appropriate temperature and time for the test organism (overnight to a week). Colonies or plaques are counted. A "positive- hole" correction is applied14 to the one-stage Andersen data to correct for undercounting at high concentrations. 3.2: Sampling Methods Requirements Inert aerosol sampling method requirements and critical dimensions and configurations of the test apparatus are specified in ASHRAE 52.2-1999. Bioaerosol sampling methodology will comply where appropriate. Bioaerosol samplers are operated according to the manufacturer's specifications. The vacuum pumps required for operating the samplers are calibrated. 3.3: Sample Handling and Custody Requirements Sampling and handling procedures will be described in testing laboratory SOPs. These SOPS will address any anticipated failures and the methods that will be employed to overcome these failures. Most of the methods are well-known sampling methods; therefore, sampling failures are not anticipated. Supporting measurements, such as temperature, relative humidity or atmospheric pressure, will be recorded in laboratory data logs, run sheets or notebooks. Upon receipt of the test devices, each will be serially numbered using a permanent marker (or other means as appropriate). All devices will be stored in a secure, temperature and humidity controlled room. 3.4: Analytical Methods Requirements The analytical method requirements for the inert aerosol testing are described in ASHRAE 52.2- 1999. A testing facility will have its own SOPs for the biological analysis. 3.5: Quality Control Requirements The apparatus will be tested to verify that the test rig and sampling procedures are capable of providing quantitatively reliable particle size measurements. Appendix A contains quality control information for inert aerosols (Table Al), the SMPSr (Table A2) and bioaerosols (Table A3). 3.6: Instrument/Equipment Testing, Inspection, and Maintenance Requirements Qualification tests will be conducted as required by the table shown in Appendix A. Typically, these tests are run as part of each test run, monthly, biannually, or after a change that may alter performance. 3.7: Instrument Calibration and Frequency Calibration will be performed in accordance with the manufacturer's recommendations or annually. Recommended instrument calibration frequencies are provided in the respective SOPs and manufacturer's manuals. ------- Research Triangle Institute Page 13 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners 3.8: Inspection/Acceptance Requirements for Supplies and Consumables Chemicals, supplies, and other consumables will be purchased from sources that have provided high quality products to the laboratory in the past. Materials such as growth media will be purchased from a single source to help ensure uniformity throughout the duration of the project. All supplies will be inspected by the lab personnel. 3.9: Data Management The work performed using this protocol will conform to the quality management plan for the APCT Center15. Guidelines for data management include the description, location, format, and organization of all types of records. The senior technical staff will oversee all data management activities. This section identifies the activities and processes planned for documenting the traceability of the data, calibrations, and information in the verification report. 3.9.1: Data Recording Data for this task will be collected either by computer or by manual (handwritten) entries. Observations and records (e.g., sample description and collection information) will be recorded manually in lab notebooks kept exclusively for this task. Output data generated by the OPC instruments will be transferred directly to a computer file and stored as a spreadsheet; printed output will be taped into the lab notebook. 3.9.2: Data Analysis Inert Aerosol Data The computation of inert aerosol filtration efficiency is based on the ratio of the downstream-to- upstream particle concentrations corrected on a channel-by-channel basis for: • Background counts (i.e., upstream and downstream counts observed when the aerosol generator is off) and • For the correlation ratio measured at the start of the test sequence. A minimum of two background and six upstream and six downstream counts will be taken. These data will be used for determining filtration efficiency by computing the observed penetration (Pobserved) (Eq. 1): (D-D,) p _ ± t> ' p -: observed /TJ TT \ 4' where: D = Downstream particle count, Db = Downstream background count, U = Upstream count, and Ub = Upstream background count. To remove system bias, the observed penetration is corrected by the correlation ratio (R) (the measured during a blank control test for which no device is installed in the duct) (Eq. 2). corrected ~ observed Eq. 2 ------- Research Triangle Institute Page 14 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners The inactivation efficiency is then computed (Eq. 3). Inactivation Efficiency (%) = 1 00 (1 - Pcorrected ) Eq . 3 Bioaerosol Data Data analysis will be performed using commercially available software (Microsoft Excel16) to enter the raw data into a spreadsheet and calculate results from a series of equations. Samples will be collected simultaneously using multiple samplers. A minimum of five, usually ten, replicates will be collected for each efficiency determination. The mean upstream and downstream CPUs will be calculated as (Eq. 4): n n _ *'=! and £> = *'=! Eq. 4 n n where: D; = Downstream count of the ith sample and n is the number of replicate samples collected and U; = Upstream count of the ith sample and n is the number of replicate samples collected. The calculation of the penetration is based on the ratio of the downstream to upstream culturable counts. The penetration with the device installed in the test rig (Pmeasured) is shown in the following equation (Eq. 5): P = - measured /U Eq. 5 where: D = Mean downstream count with a device installed in the test rig and U = Mean upstream count with a device installed in the test rig. The PIOO (no device installed in the test rig or device turned off) is calculated as the Pmeasured but using the results of the no device tests (Eq. 6). P,m - Moo/ „ , ~ 77 E. 6 ^ where: ------- Research Triangle Institute Page 15 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Dwo = Mean downstream count with no device or device turned off in the test rig and C/100 = Mean upstream count with no device or device turned off in the test rig. To remove system bias, the Pmeasured is corrected by the penetration of a blank "no device" test for which no air cleaner is installed in the duct (Pioo) (Eq. 7). P / p = measured/ corrected /P Eq. 7 / 100 H The collect!on/inactivation efficiency is then calculated as shown in Eq. 8. Collection! Inactivation Efficiency (%) = 100 (1 - Pcomcted ) Eq. 8 The precision DQO for bioaerosol inactivation efficiency will be calculated based as ± one standard deviation of penetration computed from the coefficient of variance of upstream and downstream culturable counts as shown in Eq. 9. Std. Deviation = Pmeasured (CV* + CFD2) Eq. 9 where: Pmeasured = Penetration calculated from the upstream and downstream culturable counts, CVu = Coefficient of variance for the upstream counts, and CVo = Coefficient of variance for the downstream counts. 3.9.3: Data Storage and Retrieval Laboratory notebooks containing manually recorded information and data output generated from instrumentation will be stored in the custody of the appropriate technical lead for the duration of the project. Spreadsheet files including raw and calculated data will be stored on computers. The files will be downloaded to a network server backed up nightly on magnetic tape. ETV policy requires that project files be archived offsite at a secure facility for a minimum of 7 years following the end of the project. ------- Research Triangle Institute Page 16 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners References 1. RTI. 1999. Environmental Technology Verification Test Protocol for General Ventilation Filters. Research Triangle Park, NC. http://www.epa.gov/etv/pdfs/vp/07_vp_filters.pdf 2. RTI. 1999. Environmental Technology Verification Test Plan for General Ventilation Filters. Research Triangle Park, NC. http://www.epa. gov/etv/pdfs/testplan/07_tp_093099_general.pdf 3. RTI. 2004. Environmental Technology Verification Protocol for Measuring Efficiency for Biological or Chemical Challenges of Technologies for Cleaning Building Ventilation Air. Research Triangle Park, NC. Available from RTI. 4. RTI. 2003. Environmental Technology Verification Test Plan for Biological Testing of General Ventilation Filters. Research Triangle Park, NC. http://www.epa.gov/etv/pdfs/testplan/10_tp_bio.pdf 5. RTI. 2005. Technology Testing and Evaluation Program, Test/QA Plan for Bioaerosol Inactivation Efficiency by HVAC In-Duct Ultraviolet Light Air Cleaners. Research Triangle Park, NC. Available from RTI. 6. ANSI/ASHRAE (American National Standards Institute/American Society of Heating, Refrigerating and Air-Conditioning Engineers). 1999. ANSI/ASHRAE Standard 52.2-1999, Method of Testing General Ventilation Air-Cleaning Devices for Removal Efficiency by Particle Size, Atlanta, GA. 7. Foarde, K.K. and J.T. Hanley. 2001. Determine the Efficacy of Antimicrobial Treatments of Fibrous Air Filters. ASHRAE Transactions. Volume 107, Part 1. 156-170. 8. Foarde, K.K. and J.T. Hanley. 1999. A New Laboratory Method for Measuring the Bioaerosol Filtration Efficiency of Air Cleaners. Proceedings: 1999 Air Filtration Conference: Fall Topical Conference pp. 47-54. 9. Foarde, K.K., J.T. Hanley, D.S. Ensor, and P.F. Roessler. 1999. Development of a Method for Measuring Single-Pass Bioaerosol Removal Efficiencies of a Room Air Cleaner. Aerosol Science and Technology. 30: 223-234. 10. ASTM. 2003. E178-02 Standard Practice for Dealing With Outlying Observations. American Society for Testing and Materials. West Conshohocken, PA. 11. U.S. EPA. 2002. Environmental Technology Verification Program Quality Management Plan, EPA/600/R-03/021, U.S. EPA, Cincinnati, OH, 2002. http://www.epa.gov/etv/pdfs/qmp/ETV 02 QMP.pdf 12. Knight, V. 1973. Viral and Mycoplasmal Infections of the Respiratory Tract, Lea & Febiger, Philadelphia, PA. ------- Research Triangle Institute Page 17 ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners 13. Buckland, F.E., and Tyrell, D.A.S. 1962. Loss of Infectivity on Drying Various Viruses, Nature 195:1063-1064. 14. Macher, J.M. 1989. Positive Hole Correction of Multiple-jet Impactors for Collecting Viable Microorganisms, American Industrial Hygiene Association Journal. 50: 561-568. 15.RTI (Research Triangle Institute). 2005. Quality Management Plan for Verification Testing of Air Pollution Control Technology, Revision 2.2, Research Triangle Park, NC. http://etv.rti.org/apct/pdf/apctqmp.pdf ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 18 Appendix A: Test Specifications Test specifications for the inert aerosol tests are defined in ASHRAE 52.2-1999 and shown in Table Al. These will be used for both the ASHRAE 52.2-1999 testing and the inert aerosol component of the bioaerosol test. The test specifications associated with the SMPS and the conditioning aerosol are found in Table A2. Table A3 shows the test specifications for the bioaerosol test. The test duct performance specifications applicable to all testing are found in Table A4. Test Duct /System A schematic of the test duct is shown in Figure Al. The drawing is approximately to scale. The test duct is a 610 mm (24 in.) x 610 mm (24 in.) square. The locations of the major components, including the sampling probes, test section (device holder), the aerosol generator (site of aerosol injection) are shown. There are presently no standards available to directly "calibrate" the test system for penetration. However, a number of parameters can be checked to verify proper performance. 0% and 100% penetration measurements are made by using a HEPA filter and an empty (no device) test section, respectively, using the optical particle counter (OPC) and KC1 as the inert particulate. Separate tests with the bioaerosol will be done using the test bioaerosol and the bioaerosol samplers. The flow rate will be measured via the pressure drop across an ASME long radius flow nozzle (i.e., nozzle size 8V2 in.) mounted in the center of the duct downstream of the device. It will be the primary standard for the laboratory. Prior to use, the nozzle is visually inspected to be free from defects. The installation of the nozzle in the duct will be inspected to confirm that it is seated in place. Exhaust to Room Room X Air " — ' 1 ' H_ C^K^ Y 5 A S h t \ \ In lei Blower \ BE Flow Control Valve ASME Nozzle Outlet Filter Bank 9 1 iff if c[ 11 1 0(i ^ — Filter / Upstream Mixer Ae ro so 1 ^ Generator S= Bio lo g ic a 1 Sampling Downstream Mixer P ^ = *" '^ . )J Pfvtice Eaokup Filter ; Section , ... 1 Holder(Used VMien ^ Dust-Loading) D logical mpling Figure Al. Schematic of test duct (top view) used for filter testing. Drawing is approximately to scale with the duct being 610x610 mm2 (24 x 24 in.2) and shows the location of aerosol injection, mixing baffles, test section, ASME flow nozzle, and OPC sampling probes. ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 19 The pressure drop across the device will be measured with an inclined manometer and/or a digital micromanometer. The zero and level of the manometer will be confirmed and connecting tubing inspected for integrity. Measurements of the in-duct temperature and relative humidity (RH) and room atmospheric pressure will be made. These measurements are not critical to the program and are being collected simply to document the general test environment. A wet and dry bulb psychrometer will be used for determination of temperature and relative humidity and an aneroid barometer for atmospheric pressure (periodically compared to a mercury barometer in an adjoining lab). For the bioaerosol, the RH goal is 20 - 70%. No specific quality control checks on these instruments are planned other than an inspection of the instruments for mechanical faults (e.g., mercury separation in the thermometers, poor tubing connections), and inspection of the data for reasonableness. Table Al. Quality Control Parameters For Inert Aerosol Tests Parameter Minimum counts per OPCa channel for challenge aerosol Maximum total OPC count per sample 100% Efficiency test (0% Penetration) 100% Penetration (correlation test) Penetration error limit for OPC data OPC calibration: primary calibration Frequency and description For each efficiency test, the total number of particles counted per OPC sizing channel for the upstream challenge aerosol is computed. Each efficiency test. Monthly. A HEP A filter is used for the test device. A 100% penetration test performed at least once per week during testing. Each test. Statistical check of data quality. Expected to be achieved on tests of clean air cleaners. May not always be achieved with dust-loaded air cleaners if the air cleaner sheds a significant amount of the collected dust. Primary calibration performed by manufacturer at manufacturer- specified intervals; but at least annually. Control Limits Minimum total of 500 particle counts per channel. Not to exceed maximum challenge aerosol concentration determined in the OPC upper concentration limit test referenced in Table A4. Measured penetration must be <1%. Particle Acceptable Size range Penetration Range: 0.3tol|im: 0.90 to 1.10 Ito3|im: 0.80 to 1.20 3tolO|im: 0.70 to 1.30 Per definitions and procedures of ASEQIAE 52.2 Section 10.6.46 &—;= ^ 0.07 P or 0.05 whichever is greater for 0.3 - 3 /urn v« &—j= ^ 0.15P or 0.05 whichever is greater for 3 - 5.5 /u v« T Manufacturer provides certificate of calibration. ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 20 Parameter Minimum counts per OPCa channel for challenge aerosol OPC sizing accuracy check: polystyrene latex spheres (PSL) OPC reference filter check OPC zero count Background count rate Pressure drop across empty test section Pressure drop across the air cleaner Pressure drop across the ASME flow nozzle used for measurement of airflow Aerosol charge neutralizer Filter weight Weight of ASHRAE dust fed into the test duct Frequency and description For each efficiency test, the total number of particles counted per OPC sizing channel for the upstream challenge aerosol is computed. Daily. Sample aerosolized PSL spheres. A filtration efficiency test is performed on a reference filter monthly during testing. Each correlation and initial efficiency test. Measured during correlation and clean device tests. Each correlation test. Annual. Compare to reference manometer. Annual. Compare to reference manometer. Monthly. Confirm activity of radioactive charge neutralizers. Confirm balance of corona discharge neutralizers. Filters will be weighed before and after completion of dust loading. Each test based on the change in weight of the dust on the dust- loading tray. Control Limits Minimum total of 500 particle counts per channel. Peak of distribution should be in correct OPC channel. Efficiency must be consistent with reference filter measurements made after OPC's primary calibration; efficiency within ± 10 percentage points. Less than 10 counts per sample. Upper 95% confidence limit on background counts must be less than 5% of challenge counts. Measured pressure drop must be < 0.03 in. H2O. Inclined fluid manometer or digital manometer readable to within ±0.01 in. H2O. 10% or better accuracy. Inclined fluid manometer or digital manometer readable to within ±0.01 in. H2O. 10% or better accuracy. Activity must be detected in radioactive neutralizers. Corona discharge neutralizers must be in balance. Electronic balance with 0.1 g resolution, 10% accuracy or better, calibrated annually. Electronic balance with 0.1 g resolution, 10% accuracy or better, calibrated annually. a OPC = optical particle counter ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 21 Table A2. Quality Control Parameters Associated with Scanning Mobility Particle Sizer (SMPS) and Conditioning Aerosol Parameter 0% Efficiency test (100% Penetration) SMPS: CVon upstream counts Conditioning aerosol concentration SMPS operational checks: sizing accuracy check instrument flow rates instrument zero — filtered inlet instrument zero — 0 volt setting inlet impactor photodetector Reference flow meter Reference manometer Frequency and description At least once every five efficiency tests. A 100% penetration test is performed with no device in the test section. Computed for each efficiency test based on the upstream particle counts. Measured with a condensation particle counter (CPC). At start of project and at least monthly during testing, sample aerosolized monodisperse PSL spheres. Confirmed prior to test program using reference flow meter. Checked at start of project and weekly during testing. Checked at start of project and weekly during testing. Visually confirm impactor orifice is free of debris and that the impactor plate is greased. Daily. Check at start of program. Filter on CPC inlet and/or sample pump off. Bios International Model DryCal DC1 Primary Air Flow Meter (soapless piston-type cell) or the Gilabrator (good for lower flowrates). Used to confirm SMPS flow rates at beginning of program. TSI Model 8702/8704 digital manometer and/or Meriam Model 50MH10-8 inclined fluid manometer. Control Limit Particle Acceptable Size range Penetration Range: 0.01- 1.0 |im 0.70 to 1.30 <0.30 Concentration will not exceed instrument's specified concentration limit. A relative peak in the number distribution is observed within 20% of the PSL particle diameter Flows should be within 10% of set points. < 0. 1 particle/cm3 a counted by CPC. < 0.1 particles/cm3 counted by CPC. Orifice clear of visible obstructions. Impactor has very thin film of vacuum grease. 0 ± 0.05 volts Based on the fundamental nature of this positive displacement piston flow meter, the manufacturer's accuracy claim is accepted. The unit is visually inspected for proper operation prior to use. Digital manometer to receive primary calibration by manufacturer within prior 12 months. Fluid manometer inspected for zero and level. a cm = cubic centimeter ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 22 Table A3. Quality Control Parameters for Bioaerosols Parameter Minimum upstream counts for samplers Maximum counts for samplers 100% Penetration (correlation test) Upstream CPUs Upstream PFUs Frequency and description Each efficiency test. Each efficiency test. Performed at least once per test sequence per organism Each test. Statistical check of data Each test. Statistical check of data Control Limits Minimum of 10 CFUVplate or PFUb/plate Maximum of 400 CFU/plate or 800 PFUb/plate Test Acceptable Organism Penetration Range: B.subtilis 0.85 to 1.1 5 S. epidermidis 0. 80 to 1 .20 MS2 0.80 to 1.20 A. versicolor 0.85 to 1.15 CVC < 0.20 CVC< 0.35 a CPU = colony forming units b PFU = plaque forming unit 0 CV = coefficient of variance ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Table A4. Quality Control Parameters for the Test Duct Page 23 Parameter Air velocity uniformity based on traverse measurements over a nine-point cross-sectional grid at the test flow rate. Performed upstream of the test section using a TSI Model 8345 digital thermal anemometer. Inert aerosol uniformity based on traverse measurements over a nine-point cross-sectional grid at the test flow rate. Performed upstream of the test section. Inert downstream mixing based on nine-point perimeter injection grid at the test section and center-of-duct readings at the downstream probe locations. 100% Efficiency test based on HEPA filter test. 100% Penetration (correlation test) OPCb upper concentration limit based on limiting the concentration to below the level corresponding to the onset of coincidence error. Aerosol generator response time Duct leakage Ratio of leak rate to test flow rate. Determined by sealing the duct at inlet HEPA filter bank and at the ASME flow nozzle locations followed by metering in air to achieve a steady duct pressure. The flow rate of the metering air (equal to the leakage flow) is measured for a range of duct pressures. OPC zero count check OPC sizing accuracy check based on sampling aerosolized monodisperse PSL spheres of known size. Aerosol neutralizer activity (if radioactive source is used) Dust feeder air flow rate as function of discharge pressure based on measuring the required dust feeder air gauge pressure to achieve 425 L/min. (15 cfm) airflow. Final device efficiency Based on injecting 100 g of dust and computing weight change of the filter. Control Limits CV < 10% CV<15% CV<10% Efficiency > 99% Particle Acceptable Size range Penetration Range: 0.3tol|im: 0.90 to 1.10 Ito3|im: 0.80 to 1.20 3tolO|im: 0.70 to 1.30 No predetermined level, but must be established prior to testing. No predetermined level. Ratio < 1.0% < 10 counts per sample Relative maximum must appear in the appropriate sizing channel. Radioactivity must be detected. No predetermined value. 100 ± 2 g of dust captured for 100 g injected. ' CV = coefficient of variance ' OPC = optical particle counter ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 24 Appendix B: Inert Aerosol Run Sheet Date: Test Operator: Staple photo to back of page 2. _Charge Number: Physical Description of Device: Test Requested by: Manufacturer: Product Name: Model: Condition: New or From Field No damage or Slight frame damage and/or Media damage (Circle all that apply) Other/describe damage: Product type: Other Attributes: Height Width Thickness Media Type (if applicable) Media Color (if applicable) Correlation Test: (use 3/3 -10/9 - 3/3 sampling) Date: Time: Flow rate manometer zeroed and level: Device pressure drop manometer zeroed and level: OPC clock correct: Valve switch on: OPC: (Set for 0.10 ft3 samples with 15 second purge; use 3/3 -10/9 - 3/3 sampling) 20 min warm up / Flow is 0.25 cfm / CI-226 switch "Low" / or n/a Zero Check < 10 total / sample enter actual count HEPA capsule or In-duct Daily PSL check (Enter size when performed or / if done earlier today) File Name c:\climet\rpmmddyyss RP MM DD YY SS must meet <10 criteria at least once per ASHRAE 52.2 test. Notify project manager if limit is exceeded. ** Save daily check to disk using file name. RPMMDDYY-HEPA-PSL.TXT ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Test Conditions: Page 25 Flow rate (cfm)++ Flow Manometer (inchH2O) Dry Bulb Temp. (F) Limits = 50-100 °F Wet Bulb Temp. (°F) RH Limits = 20-65% Arm Pressure (inches Hg) XX. XX Is flowrate MERV eligible for this size device? Y or N (see page 3) Aerosol Generator: JNo-Device Pressure Dro Aerosol Type KCL Pump setting 1.2cc/ min Drying Air 4 cfm 240 cfh Nozzle air pressure (psi) Nozzle air flowmt. p Upper Concentration Target * enter Ch 1 count Lower Concentration target ** enter Ch 1 5 cnt At start of test x.xx(x) must be < 0.03" At end of test x.xx(x) must be < 0.03" Channel 1 targets: CI-500 = 3,000; Cl-Spectro = 3,000; CI-226 = 5,000 Channel 15 target: 72 -150 counts per sample desired range. Notify project manager if these targets are not met. Using "Correlation" graph in spreadsheet, does data look reasonable? yes no (Should be near 1.0) ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Appendix C: Bioaerosol Run Sheet Page 26 Device #:_ Date: Run#: Test Operator: Climet Filename: Test Conditions: Test Flow Rate CFM ASME Nozzle Pressure Drop in. H2O Temperature °F RH % Ambient Atm Pressure in. Hg Biological Suspension: Organism: Suspension Prep:_ Drying Air: Nebulizer Pressure: Initial Volume: Time On: Biological Sampling: Sample # Time Run Begins Sample Length (min.) Media Ul, U2, U3 D1,D2, D3 D4, D5, D6 U4, U5, U6 U7, U8, U9 ------- Research Triangle Institute ETV Protocol Bioaerosol and Aerosol Testing of General Ventilation Air Cleaners Page 27 D7, D8, D9 D10,D11,D12 U10,U11,U12 U; = upstream sample ;' D; = downstream sample ;' Rotometer/Vac #1 Reading: Rotometer/Vac #2 Reading: Rotometer/Vac #3 Reading: ------- |