EPA 600/R-11/143 | November 2011 | www.epa.gov/ord
United States
Environmental Protection
Agency
    ] National Institute for
    f Occupational Safety and Health
             Surface Analysis Using
             Wipes for the Determination
             of Nitrogen Mustard
             Degradation  Products by Liquid
             Chromatography/Tandem Mass
             Spectrometry (LC/MS/MS)
             SAMPLING AND ANALYTICAL
             PROCEDURE FOR ANALYSIS OF
             SURFACES USING WIPES
             REVISION 2
Office of Research and Development
National Homeland Security Research Center
Centers for Disease Control and Prevention
National Institute for Occupational Safety and Health

-------
                                                                   EPA/600/R-11/143
SURFACE ANALYSIS USING WIPES FOR THE DETERMINATION OF NITROGEN
               MUSTARD DEGRADATION PRODUCTS BY LIQUID
      CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY (LC/MS/MS)
            Sampling and Analytical Procedure for Analysis of Surfaces Using Wipes
                                      Revision 2
                       United States Environmental Protection Agency
                       National Homeland Security Research Center
                            26 W. Martin Luther King Jr. Drive
                                 Cincinnati, OH 45268

                                         and

                        Centers for Disease Control and Prevention
                    National Institute for Occupational Safety and Health
                                   5555 Ridge Ave
                                 Cincinnati, OH 45213

                                   Last Revised: 3/11

-------
                                                                         NHSRC/NIOSH SAP

                                                                        Revision Date: 3/30/11
                                                                               Page ii of 38
                                       DISCLAIMER
The information in this document is a single-laboratory-developed sampling and analytical procedure
(SAP) that has been funded wholly or in part by the U.S. Environmental Protection Agency (EPA) and in
collaboration with the National Homeland Security Research Center, part of EPA's Office of Research
and Development, and the National Institute of Occupational Safety and Health (NIOSH), a division of
the U.S.  Department of Health and Human  Services (DHHS),  under IA #DW-75-922440001-0.  The
method development and  document preparation were supported under contract number EP08C000010.
This document has been subjected to the Agency's review and has been approved for publication. Note
that approval does not signify that the contents necessarily reflect the views  of the Agency.  NIOSH and
EPA do not endorse the purchase or sale of any commercial products or services.

Questions concerning this document or its application should be addressed to:

Stuart Willison, Ph.D.
Project Officer
U.S. Environmental Protection Agency
National Homeland Security Research Center
26 W. Martin Luther King Drive, MS NG16 Cincinnati, OH 45268
513-569-7253
Willison.Stuart@epa.gov

Robert Streicher, Ph.D.
Project Officer
National Institute for Occupational Safety and Health Laboratories
Alice Hamilton Laboratory
5555 Ridge Avenue
Cincinnati, OH 45213
513-841-4296
Rps3@cdc.gov

-------
                                                                           NHSRC/NIOSH SAP

                                                                          Revision Date: 3/30/11
                                                                                 Page iii of 38
                                        FOREWORD
The National Homeland Security Research Center (NHSRC), part of U.S. EPA's Office of Research and
Development (ORD), is focused on developing and delivering scientifically sound, reliable, and
responsive products. These products are designed to address homeland security information gaps and
research needs that support the Agency's mission of protecting public health and the environment. A
portion of NHSRC's research is directed at decontamination of indoor surfaces, outdoor areas, and water
infrastructure. This research is conducted as part of EPA's response to chemical, biological, and
radiological (CBR) contamination incidents. NHSRC has been charged with delivering tools and
methodologies (e.g. sampling and analytical methods, sample collection protocols) that enable the rapid
characterization of indoor and outdoor areas, and water systems following terrorist attacks, and more
broadly, natural and manmade disasters.

The Selected Analytical Methods for Environmental Remediation and Recovery  (SAM) document  is a
compendium of methods that informs sample collection and analysis during the response to an incident.
SAM can be used by public and private  laboratories which are analyzing a large number of samples
associated with chemical, biological, or radiological contamination. Even though some of the  analytes in
SAM already have existing analytical methods,  others are in need  of improvements that enhance
analytical capability and meet more rigorous performance criteria.  Furthermore, not all  of the analytical
methods  listed in the SAM document address  all  possible matrices (e.g.,  water,  soil,  air,  glass)
encountered in sample collection following an incident.  The analytical methods in SAM  have been
verified in a single laboratory, but most still need to undergo multi-laboratory validation with respect to a
specific contaminant in association with a specific matrix.

The single-laboratory-developed Sampling and Analytical Procedure, described herein, demonstrates the
procedure for analysis of nitrogen  mustard  degradation products on surfaces using wipes by liquid
chromatography/tandem  mass  spectrometry  (LC/MS/MS).  A companion study report (Companion
Document for Surface Analysis Using Wipes for the Determination of Nitrogen Mustard Degradation
Products  by  Liquid Chromatography/Tandem  Mass  Spectrometry  (LC/MS/MS)),  describes  the
development efforts, a synopsis  of the supporting data collected,  and scientific justifications for the
decisions made.  NHSRC  welcomes your comments as  we move one  step closer to achieving our
homeland security mission, and our overall mission of protecting human health and the environment.
Jon Herrmann,
Director National Homeland Security Research Center

-------
                                                                       NHSRC/NIOSH SAP

                                                                      Revision Date: 3/30/11
                                                                             Page iv of 38
                                 ACKNOWLEDGMENTS
Contributions of the following individuals and organizations to the development and review of this
method are acknowledged.

United States Environmental Protection Agency (EPA)
       Office of Research and Development, National Homeland Security Research Center
       Stuart Willison, Project Officer and Method Development
       Erin Silvestri, Project Officer

       EPA Regions
       Lawrence Zintek (Region 5), Technical Reviewer
       Lukas Oudejans (RTP), Technical Reviewer
       Lawrence Kaelin (OSWER), Technical Reviewer

Centers for Disease Control and Prevention
       National Institute for Occupational Safety and Health
       Jack Pretty
       Robert Streicher
                                            IV

-------
                                                                            NHSRC/NIOSH SAP

                                                                           Revision Date: 3/30/11
                                                                                  Page v of 38
                                   EXECUTIVE SUMMARY

Wipe sampling can be performed quickly and easily when direct extraction is not always possible (e.g.,
walls, floors and furniture). As a result, wipe sampling is preferred for analysis without destruction of the
tested surface.  However, wipe sampling can remove analytes only from the surface of a material, which
could result in lower  recoveries  and produce less reliable quantitative data from porous surfaces.  It is
therefore important to understand  wipe efficiencies and the materials being wiped.  This  procedure
assesses the recoveries from various porous and nonporous surfaces to determine the presence of nitrogen
mustard degradation products.  Wipes were analyzed using 100% methanol extraction  by sonication,
filtration,   and  concentration   followed  by   analysis  by   liquid  chromatography   electrospray
ionization/tandem mass spectrometry (LC/ESI-MS/MS) by direct injection without derivatization.  Data
generated from a formica surface resulted in detection limits of 0.12 |ig/cm2 for TEA, 0.06 |ig/cm2 for
EDEA, 0.07 |ig/cm2 for MDEA, and 0.04 |ig/cm2 for DEA. Accuracy and precision data were generated
from each tested surface fortified with these analytes, then qualitatively and quantitatively determined.

-------
                                                            NHSRC/NIOSH SAP

                                                           Revision Date: 3/30/11
                                                                 Page vi of 38
 SURFACE ANALYSIS USING WIPES FOR THE DETERMINATION OF NITROGEN MUSTARD
      DEGRADATION PRODUCTS BY LIQUID CHROMATOGRAPHY/TANDEM MASS
                          SPECTROMETRY (LC/MS/MS)

                            TABLE OF CONTENTS
      SECTION                                                    PGNO.
      DISCLAIMER                                                i

      FOREWORD                                                 ii

      ACKNOWLEDGMENTS                                         iii

      EXECUTIVE SUMMARY                                        v

      LIST OF TABLES                                              vii

      LIST OF ACRONYMS AND ABBREVIATIONS                          viii

1.     SCOPE AND APPLICATION                                      1

2.     SUMMARY OF METHOD                                        2

3.     DEFINITIONS                                                3

4.     INTERFERENCES                                              4

5.     HEALTH AND SAFETY                                         5

6.     EQUIPMENT AND SUPPLIES                                     5

7.     REAGENTS AND STANDARDS                                    7

8.     SAMPLE COLLECTION, PRESERVATION AND STORAGE                 8

9.     QUALITY CONTROL                                           9

10.    INSTRUMENT CALIBRATION AND STANDARDIZATION                 13

11.    ANALYTICAL PROCEDURE                                      14

12.    DATA ANALYSIS AND CALCULATIONS                             16

13.    METHOD PERFORMANCE                                       17

14.    POLLUTION PREVENTION                                      18

15.    WASTE MANAGEMENT                                        18

16.    REFERENCES                                                18

17.    TABLES AND VALIDATION DATA                                 20

18.    ATTACHMENTS                                              25
                                     VI

-------
                                                                        NHSRC/NIOSH SAP

                                                                       Revision Date: 3/30/11
                                                                              Page vii of 38
                                     LIST OF TABLES


Table 1. Method Parameters	21
Table 2. Holding Time Sample Stability of Nitrogen Mustard Degradation Analytes	21
Table 3. Concentrations of Calibration Standards  (ng/mL)	22
Table 4. MRM Ion Transitions, Retention Time and Variable Mass Spectrometer Parameters	22
Table 5. Gradient Conditions for Liquid Chromatography	23
Table 6. ESI+-MS/MS Conditions	23
Table 7. Materials Tested for the Wipe Analysis of Nitrogen Mustard Degradation Products	24
Table 8. List of Consumable Materials Used During Sampling	24
                                            VII

-------
LIST OF ACRONYMS AND ABBREVIATIONS

ACN     Acetonitrile
AS       Analyte Standard
CAL     Calibration Standard
CAS      Chemical Abstracts Service
CBR     Chemical, Biological and/or Radiological
CCC     Continuing Calibration Check
CID      Collisionally Induced Dissociation
CV       Calibration Verification
CWA     Chemical Warfare Agent
DBA     Diethanolamine
DHHS     U.S. Department of Health and Human Services
DL       Detection limit
EDEA     TV-Ethyldiethanolamine
EPA      U.S. Environmental Protection Agency
ERLN     Environmental Response Laboratory Network
ESI       Electrospray lonization
IDC      Initial Demonstration of Capability
IDL      Instrument Detection Limit
LC       Liquid Chromatography
LC/MS/MS Liquid Chromatography Coupled with Tandem Mass Spectrometry
LFB      Laboratory Fortified Blank
LFSM     Laboratory Fortified Sample Matrix
LFSMD   Laboratory Fortified Sample Matrix Duplicate
LRB      Laboratory Reagent Blank
MDEA    TV-Methyldiethanolamine
MDL     Method Detection Limit
MRL     Minimum Reporting Limit
MRM     Multiple Reaction Monitoring
MS       Mass Spectrometry
MSDS     Material Safety Data Sheet
MS/MS   Tandem Mass Spectrometry
MSP     Method Specific Parameter
NFL^OAc  Ammonium Acetate
NHSRC   National Homeland Security Research Center
NIOSH   National Institute for Occupational Safety and Health
NIST     National Institute of Standards and Technology
ORD     U.S. EPA's Office of Research and Development
OSHA     Occupational Safety and Health Administration
PPB      Parts per Billion
PPM     Parts per Million
P&A     Precision and Accuracy
PVDF     Polyvinylidene Fluoride
QA       Quality Assurance
QC       Quality Control
QL       Quantitation Limit
QMP     Quality Management Plan
REC      Percent Recovery
RL       Reporting Limit
                                                                    NHSRC/NIOSH SAP

                                                                   Revision Date: 3/30/11
                                                                          Page viii of 38
                                        VIM

-------
                                                                    NHSRC/NIOSH SAP

                                                                   Revision Date: 3/30/11
                                                                          Page ix of 38
RSD      Relative Standard Deviation
RT       Retention Time
RTS      Retention Time Shift
SAM     Standardized Analytical Methods for Environmental Restoration Following Homeland
          Security Events
SAP      Sampling and Analytical Procedure
SD       Standard Deviation
S/N       Signal to Noise
SS        Surrogate Standard
SSS       Stock Standard Solution
TC       Target Compound
TEA      Triethanolamine
UPLC     Ultra-Performance Liquid Chromatography
X         Average Percent Recovery
                                         IX

-------
                                                                            NHSRC/NIOSH SAP

                                                                           Revision Date: 3/30/11
                                                                                  Page 1 of 38
1.   SCOPE AND APPLICATION

       1.1.     This  procedure  covers the determination  of nitrogen mustard degradation  products on
               surfaces using wipes.  Surfaces were wiped  and wipes were analyzed using  100%
               methanol extraction by sonication, filtration, and concentration followed by analysis by
               liquid chromatography electrospray ionization/tandem  mass  spectrometry  (LC/ESI-
               MS/MS) by direct injection without  derivatization.  Detection limit data were generated
               for all analytes of interest on surfaces.  Accuracy and precision data were generated from
               each   surface  fortified with these  analytes,  then qualitatively  and  quantitatively
               determined. The following analytes were determined using this procedure:

                          Analyte                              CAS Registry Number
                         Triethanolamine (TEA)                       102-71-6
                         7V-Ethyldiethanolamine (EDEA)               139-87-7
                         7V-Methyldiethanolamine  (MDEA)             105-59-9
                         Diethanolamine (DEA)                       111 -42-2

       1.2.     Wipe sampling can be performed quickly and easily when direct extraction is not always
               possible (e.g., walls,  floors  and  furniture).  As a result, wipe  sampling is preferred for
               analysis without destruction of the tested surface. However, wipe sampling can remove
               analyte only from the surface of a material, which could result in lower recoveries and
               produce less reliable quantitative data from porous surfaces.  It is therefore  important to
               understand wipe efficiencies and the  materials being wiped.  This procedure assesses the
               recoveries from various porous and nonporous surfaces using  wipes.

       1.3.     Detection limit  (DL)  metrics were presented using EPA conventions1"3.  The detection
               limit  was defined  as the statistically calculated minimum  concentration  that can be
               measured with  99%  confidence that the reported  value is  greater than  zero.   The
               statistical procedure, utilizing the Laboratory Fortified Sample Matrix samples (LFSM)
               and their duplicates, will be used to calculate uncertainty.  Precision and accuracy (P&A)
               studies should be performed as an initial demonstration of capability (IDC) and after any
               modifications to the  procedure, including  changes in instrumentation  and  operating
               conditions.   These studies  will evaluate  whether the reporting  limits and calibration
               standard concentrations are appropriate.

       1.4.     The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets data
               quality objectives that are developed based on the  intended use of this sampling and
               analytical procedure (SAP).  The MRL is the  lowest true concentration for which the
               future recovery  is predicted (between 50 and 150% recovery) and is listed as the lowest
               calibration level (Level 1).

       1.5.     This  method was intended  for use by analysts  skilled in the  operation of LC/MS/MS
               instruments  and the  interpretation   of the associated  data.    Due  to the  inherent
               complexities of LC/MS/MS analysis, including the need to relate sample characteristics
               to  analytical  performance,  laboratories  should  update  their initial  estimates  of
               performance and should strive to tighten their quality control  limits as more experience is
               gained with this particular procedure.

-------
                                                                            NHSRC/NIOSH SAP

                                                                           Revision Date: 3/30/11
                                                                                  Page 2 of 38
        1.6.    METHOD FLEXIBILITY

               Many variants of liquid chromatography (LC) and Tandem Mass Spectrometry (MS/MS)
               technology are currently in operation.  In addition, variability exists  in the sources of
               wipe materials, wipe composition, and compatibility of various wipe materials with some
               surfaces.   This  procedure  was developed using a triple quadrupole LC/MS/MS, with
               optimized LC conditions and wipe materials. The procedure has been verified using only
               the specified equipment and conditions. Other types of LC/MS/MS instrumentation, LC
               conditions, and wipe/collection materials can be used for analysis as long as similar
               performance is demonstrated and the quality control measures can be observed.
2.  SUMMARY OF METHOD

       2.1.    Samples are collected from surfaces with wipes and stored at 4 °C (± 2 °C) for samples
               not immediately analyzed within a 24-hour time period.  When the samples are ready to
               be analyzed, samples are spiked with a surrogate compound, the appropriate solvent is
               added, the sample solution is sonicated, the solution is extracted with a syringe filter unit,
               then the extract is concentrated and analyzed directly by LC/MS/MS operated in  the
               positive electrospray ionization (ESI+) mode.

       2.2.    Each target compound was separated and identified by retention time and by comparing
               the sample primary multiple reaction monitoring (MRM) transition to the known standard
               MRM transition  from reference  spectra under identical LC/MS/MS  conditions.  The
               retention  time  for the analytes  of interest must fall  within the retention time  of the
               standard  (within ± 5%).  The concentration  of each analyte  is  determined by  the
               instrumentation software using external calibration.

       2.3.    The detection  limit  (DL) and quantitation limit (QL)  for these compounds were
               calculated using an EPA approach and are  listed in Table 1 and Attachment 18.1.  The
               precision   and  accuracy  (P&A)  quality  control acceptance  criteria  are  shown  in
               Attachment 18.2.  Stability studies suggest samples can be stored up to 28 days (Table 2)
               at 4 °C (± 2 °C). The concentrations of the  calibration standards are listed in Table 3 and
               the retention times, mass transitions, and mass spectrometer parameters are listed in
               Table 4. The gradient conditions and ESI-MS/MS conditions are listed in Tables 5 and 6,
               respectively. This SAP was tested on several wipes in previous studies to establish that
               filter paper provided the highest recoveries with the least interference for any targeted
               analytes.  Analytes spiked onto surfaces were wiped and the recoveries from both porous
               and nonporous  surfaces were reported.

               The overall performance of the filter paper, in terms of analyte recovery and fewest
               interferences, suggested that the filter paper was an appropriate wipe for recovering the
               analytes from a surface while analyzing the described nitrogen mustard degradation
               products.  Other wipes such as cotton gauze would require a pre-cleaning step due to the
               presence of interferences to the targeted analytes.

-------
                                                                      NHSRC/NIOSH SAP

                                                                     Revision Date: 3/30/11
                                                                           Page 3 of 38
3.  DEFINITIONS

       3.1    ANALYSIS BATCH - A set of samples analyzed on the same instrument, not
             exceeding a 24-hour period and including no more than 20 field samples, beginning
             and ending with the analysis of the appropriate continuing calibration check (CCC)
             standards. Additional CCCs may be required depending on the number of samples in
             the analysis batch and/or the number of field samples.

       3.2    CALIBRATION STANDARD (CAL) - A solution prepared from the analyte stock
             standard solution and the internal standard. The CAL solutions are used to calibrate
             the instrument response with respect to analyte concentration.

       3.3    COLLISIONALLY INDUCED DISSOCIATION (CID) - The process of converting
             the precursor ion's translational energy into internal energy by collisions with neutral
             gas molecules to bring about dissociation into product ions.

       3.4    CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing
             the method analytes and surrogate standard. The CCC is analyzed periodically to
             verify the accuracy of the existing calibration for those analytes at or near the mid-
             level concentrations.

       3.5    DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be
             identified, measured, and reported with 99% confidence that the analyte
             concentration is greater than zero.

       3.6    SURROGATE STANDARD (SS) - A pure chemical added to a standard solution in
             a known amount(s) and used to measure the relative response of other method
             analytes that are components of the same solution. The surrogate standard must be a
             chemical that is structurally similar to the method analytes, has no potential to be
             present in samples, and is not a method analyte. This method uses a deuterated
             analyte.

       3.8    LABORATORY FORTIFIED  BLANK (LFB) - A volume of solvent or other blank
             matrix to which known quantities of the method analytes are added in the laboratory.
             The LFB is analyzed exactly like a sample, and its purpose is to demonstrate that the
             methodology is in control and that the laboratory is capable of making accurate  and
             precise measurements.

       3.9    LABORATORY FORTIFIED  SAMPLE MATRIX (LFSM) - A preserved field
             sample to which known quantities of the method analytes are added in the laboratory.
             The LFSM is processed and analyzed exactly like a sample, and its purpose is to
             determine whether the sample matrix contributes bias to the analytical results. The
             background concentrations of the analytes in the sample matrix must be determined
             in a separate sample and the measured values in the LFSM must be corrected for
             background concentrations.

-------
                                                                       NHSRC/NIOSH SAP

                                                                      Revision Date: 3/30/11
                                                                             Page 4 of 38
       3.10   LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A
              duplicate of the field sample used to prepare the LFSM. The LFSMD is fortified and
              analyzed identically to the LFSM. The LFSMD is used to assess method precision
              when the occurrence of method analytes is low.

       3.11   LABORATORY REAGENT BLANK (LRB) - An aliquot of solvent or other blank
              matrix that is treated exactly as a sample  including exposure to all glassware,
              equipment,  solvents and reagents and surrogate standards that are used in the analysis
              batch. The LRB is used to determine if method analytes or other interferences are
              present in the laboratory environment, the reagents, or the apparatus.

       3.12   MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by
              vendors concerning a chemical's toxicity, health hazards, physical properties, fire,
              and reactivity data including storage, spill, and handling precautions.

       3.13   MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can  be
              reported as  a quantitated value for a method analyte in a sample following analysis.
              This defined concentration can be no lower than the concentration of the lowest
              calibration standard for that analyte and can be used only if acceptable quality control
              (QC) criteria for this standard are met. A procedure for verifying a laboratory's MRL
              is provided  in Section 9.2.4.

       3.14   PRECURSOR  ION - For the purpose of this method, the precursor ion is the
              protonated molecule ([M+H]+) or adduct ion of the method analyte. In MS/MS, the
              precursor ion is mass-selected and fragmented by collisionally induced dissociation
              (CID) to produce distinctive product ions of lower mass.

       3.15   PRODUCT ION - For the purpose of this method, a product ion is one of the
              fragment ions produced in MS/MS by CID of the precursor ion.

       3.16   STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing one
              or more method analytes  prepared in the laboratory using assayed reference materials
              or purchased from a reputable commercial source.

4.  INTERFERENCES

   Procedural interferences may be caused by contaminants in solvents, reagents,  glassware and other
   apparatus that lead to discrete artifacts or elevated baselines in the  selected ion current profiles.  All
   of these materials must  routinely be demonstrated to be free from  interferences by analyzing LRBs
   (Section 9.4.1) under the same  conditions as the samples.4  Subtraction of blank values from sample
   results is not performed.

       4.1     All  reagents and solvents should be of pesticide grade purity or higher to minimize
              interference problems. All glassware should be cleaned and demonstrate to be free from
              interferences.

       4.2     Matrix interferences may be caused by contaminants from the sample, sampling devices
              or storage containers. The extent of matrix interferences will vary considerably from

-------
                                                                          NHSRC/NIOSH SAP

                                                                         Revision Date: 3/30/11
                                                                                Page 5 of 38
              sample source to sample source, depending upon variations in the sample matrix. Wipe
              matrix interferences and contaminants are likely to be present and may have an effect on
              the recoveries for the analytical procedure. These interferences lead to elevated baselines
              and artifacts that may be interpreted as positives.

       4.3    Matrix effects are well known phenomena of ESI-MS techniques, especially for co-
              eluting compounds.  Managing the unpredictable  suppression and enhancement caused
              by these effects is recognized as an integral part of the performance and verification of an
              ESI-MS procedure.  The data presented in this procedure were designed to demonstrate
              that the procedure is capable  of functioning  with realistic  samples.  Each  analyst is
              encouraged to observe appropriate precautions and follow the described QC procedures
              to help minimize the influence of ESI-MS  matrix effects on the  data reported.  Matrix
              effects include ion suppression/enhancement, high background and improper ion ratios.

5.   HEALTH AND SAFETY

    The toxicity and carcinogenicity of each reagent used in this method have not been defined precisely.
    However, each chemical compound was treated as a health hazard.  Exposure  to these  chemicals
    should be reduced to the lowest possible level and proper protective equipment should be worn for
    skin,  eyes, etc. Each laboratory is responsible for maintaining an awareness of Occupational Safety
    and Health Administration (OSHA) regulations regarding safe handling of chemicals  used in this
    method.  A reference file of material safety data sheets (MSDSs) that address the safe handling of the
    chemicals should be made available to all personnel involved in the chemical analyses. Additional
    references are available.5"7

6.   EQUIPMENT AND SUPPLIES

    References to specific brands of equipment and catalog numbers were provided  solely  as examples
    and do not constitute an endorsement of the use of such products or suppliers.

       6.1    LC/MS/MS APPARATUS

           6.1.1    LIQUID CHROMATOGRAPH (LC)  SYSTEM - An analytical system  complete
                   with  a  temperature  programmable liquid chromatograph with a  solvent mixer
                   (Waters - Acquity™ or equivalent able to perform the analyses as described) and all
                   required accessories including syringes, solvent degasser, and autosampler.

           6.1.2   ANALYTICAL COLUMN - Waters - Atlantis™ HILIC Silica, 100 mm x 2.1 mm, 3
                  |im particle size, or equivalent.

           6.1.3   TANDEM MASS SPECTROMETER (MS/MS) SYSTEM - A MS/MS instrument,
                  Waters TQD™, or similar instrument, can be used for analysis of the target analytes.
                  A mass spectrometer capable of MRM analysis with the capability to obtain at least
                  10 scans over a peak with adequate sensitivity is required.

           6.1.4   DATA SYSTEM - MassLynx™ software (or similar software) interfaced to the
                  LC/MS  that allows the  continuous acquisition  and storage  on machine-readable
                  media of all  mass spectra obtained throughout the duration of the chromatographic
                  program.  QuanLynx™ (or similar software) is used for all quantitation for data
                  generated from the LC/MS unit.

-------
                                                                NHSRC/NIOSH SAP

                                                               Revision Date: 3/30/11
                                                                     Page 6 of 38
6.2     NITROGEN EVAPORATOR

   6.2.1   A nitrogen evaporation device, such as the N-Evap 24 - port device (Organomation
          Associates, Inc.) equipped with a water bath that can be maintained at 50 °C for final
          analyte concentration (< 10 mL volume). Evaporation times are expected to increase
          without this feature. Other nitrogen devices, which do not have a temperature control
          feature, may be used as long as QC criteria can still be reached.

6.3     EXTRACTION DEVICE

   6.3.1   SONICATOR (Fisher Scientific Catalog #: 15-335-112) or equivalent.

6.4     GLASSWARE AND MISCELLANEOUS SUPPLIES

   6.4.1   AUTOSAMPLER VIALS - Amber 2-mL autosampler vials with TeflonD-lined
          screw tops (Waters Corp., Milford, MA), or equivalent.

   6.4.2   DISPOSABLE STERILE SYRINGES - 10.0 mL ± 1% accuracy (Fisher Scientific,
          Pittsburgh, PA), or equivalent.

   6.4.3   AUTO PIPETTES - 10.0 mL, 1000 uL, 100 uL and 10 uL ± 1% accuracy.

   6.4.4   DESOLVATION GAS - Ultra Pure nitrogen gas generator or equivalent nitrogen gas
          supply. Aids in the generation of an aerosol of the ESI liquid spray and should meet
          or exceed instrument manufacturer's specifications.

   6.4.5   COLLISION GAS - Ultra  Pure  Argon gas  used in the collision cell in MS/MS
          instruments  and and should meet or exceed instrument manufacturer's specifications.

   6.4.6   ANALYTICAL BALANCE - accurate to 0.1 mg;  reference weights traceable to
          Class S or S-l weights.

   6.4.7   National Institute of Standards and Technology (NIST)-traceable thermometer.

   6.4.8   STANDARD SOLUTION FLASKS - Class A volumetric glassware

   6.4.9   SYRINGE  FILTER  - Millex®  GV Syringe-driven  filter  unit (PVDF)  0.22 |im
          (Millipore Corporation, Catalog # SLGV013NL).

   6.4.10  WIPES - Whatman 42 ashless, 55 mm filter paper (Fisher Scientific, Pittsburgh, PA,
          Catalog # 09-845A).

   6.4.11  SAMPLE COLLECTION CONTAINERS -  Clean Nalgene containers with  screw
          cap (Fisher Scientific, Pittsburgh, PA, Catalog # 11-815-10C), or equivalent.

   6.4.12  SAMPLE CONCENTRATION CONTAINERS - Sterile 15 mL conical tubes (Fisher
          Scientific, Pittsburgh, PA, Catalog # 05-538-59A), or equivalent.

-------
                                                                         NHSRC/NIOSH SAP

                                                                        Revision Date: 3/30/11
                                                                               Page 7 of 38
7.  REAGENTS AND STANDARDS

       7.1     REAGENTS AND STANDARDS

              When compound purity is assayed to be 98% or greater, the weight may be used without
              correction to  calculate the concentration of the  stock standard.  Expiration times of
              prepared  solutions  are suggested below, but  laboratories should follow  standard QC
              procedures to determine when the standards should be replaced.  Label all standards and
              verify the correct grade of solvents.  Traceability of standards is established by the
              manufacturer's specifications provided at time of purchase.

           7.1.1   SOLVENTS - Acetonitrile (CAS # 75-05-8), Methanol (CAS # 67-56-1), and Water
                  (CAS  #  7732-18-5), HPLC mass  spectrometry  pesticide  grade  or equivalent,
                  demonstrated to be free of analytes and interferences.

           7.1.2   AMMONIUM ACETATE (CAS #  631-61-8, ACS Reagent  Grade or equivalent
                  demonstrated to be free of analytes and interferences.)

           7.1.3   ACETIC ACID (CAS # 64-19-7, Concentrated, ACS Reagent Grade or equivalent
                  demonstrated to be free of analytes and interferences.)

           7.1.4   MOBILE PHASE A  - Solution A consisted of 95% of 25 mM ammonium acetate at
                  pH 4.2, and 5% of acetonitrile to prevent microbial growth. To prepare  1 L, add 1.93
                  g of ammonium acetate to water, adjust to pH 4.2 with acetic acid and dilute to 1 L
                  mark.  Add 950 mL of the 25 mM ammonium acetate  at pH 4.2 solution to  a 1L
                  container. Add  50 mL of acetonitrile. This solvent system is still  prone  to  some
                  microbial growth and should be replaced once a week.

           7.1.5   MOBILE PHASE B- Solution B was comprised of 95% acetonitrile and 5% 25 mM
                  ammonium acetate.  To prepare 1 L, add 1.93 g of ammonium acetate to 1 L of water.
                  Add 950 mL of acetonitrile to a 1L container.  Add 50 mL of the 25 mM ammonium
                  acetate solution.

           7.1.6   TARGET ANALYTES - Triethanolamine (CAS # 102-71- 6), 7V-ethyldiethanolamine
                  (CAS  # 139-87-7), 7V-methyldiethanolamine (CAS # 105-59-9) and diethanolamine
                  (CAS# 111-42-2).

           7.1.7   SURROGATE  ANALYTE-  Bis(2-hydroxyethyl)-d8-amine  (Diethanolamine-d8)
                  (CAS# 103691-51-6)

       7.2     STANDARD SOLUTIONS

              When compound purity is assayed to be at least 98% or greater, the weight can be used
              without correction to calculate the concentration of the stock standard.  Stock standards
              and all subsequent solutions should be replaced when analyzed solution concentrations
              deviate more than ± 20% from the prepared concentration.  Standards are stored protected
              from light (amber vials) and at 4 °C (± 2 °C).  Standards were estimated to be stable for
              at least a month. Although stability times are suggested, laboratories should utilize QC

-------
                                                                        NHSRC/NIOSH SAP

                                                                       Revision Date: 3/30/11
                                                                              Page 8 of 38
              practices to determine when standards should be replaced.

           7.2.1   SURROGATE STOCK STANDARD SOLUTION (SSS) (10-1000 |ig/mL)

                  A standard solution may be prepared from  a certified commercially available neat
                  compound. Isotopically-labeled surrogate, diethanolamine-d8 (CAS # 103691-51-6),
                  was obtained from CDN Isotopes.  The  surrogate was added to a 50 mL volumetric
                  flask in order to achieve a concentration of 1000 |ig/mL in solution (i.e., 50 mg or
                  45.8 |iL of diethanolamine-d8 was added to a 50 mL volumetric flask and diluted to
                  mark with methanol).  Further dilutions of the 1000 |ig/mL concentration were used
                  to obtain 100 and 10 |ig/mL solutions in methanol. Surrogate stock standard solutions
                  were stable for at least a month when stored at 4 °C.

                  (NOTE:  Although  diethanolamine-d8  was used  as  a surrogate  in this  SAP,
                  diethanolamine-d8 could be used  as  an internal  standard  for diethanolamine  for
                  quantitation purposes.  However, further evaluation would be necessary to ensure that
                  diethanolamine-dg is a viable internal standard and meets QC requirements.)

           7.2.2   ANALYTE STOCK STANDARD SOLUTION  (AS)

                  Standard solutions may be prepared from certified,  commercially  available neat
                  compounds.  All neat compounds are viscous liquids  at room temperature. Neat
                  materials of triethanolamine and diethanolamine were obtained from Chem Service
                  (West  Chester,  PA).  7V-ethyldiethanolamine  and 7V-methyldiethanolamine  were
                  obtained from Aldrich as neat materials. A standard solution concentration of 1000
                  |lg/mL for each compound was obtained in 50 mL volumetric flasks (e.g., 44.4  |iL of
                  TEA, 45.87 |iL  of DEA,  49.31  |iL of EDEA  and 48.08 |iL of MDEA were each
                  added to separate 50 mL volumetric flasks and  diluted to the mark with methanol) .
                  Further dilutions of the 1000 |ig/mL concentration can be used to obtain 100 and 10
                  |lg/mL solutions in methanol. The calibration standards were made  from appropriate
                  dilution concentration of these stock standards.

                  (NOTE: All spiking solutions should be within ten times the  DL).

           7.2.3   CALIBRATION STANDARD SOLUTION (CAL)

                  A calibration stock standard  solution (Level  7)  was prepared from the  Analyte
                  Standard   (AS)   solution  concentrations,  containing,   triethanolamine,  7V-
                  ethyldiethanolamine, 7V-methyldiethanolamine,  diethanolamine  and  the surrogate
                  diethanolamine-d8 in methanol (i.e., 250 |iL of TEA, DEA, EDEA, MDEA and DEA-
                  ds of a 100 |ig/mL solution was added to a 50 mL volumetric flask and diluted to
                  mark with methanol).   From Level 7, further dilutions were performed to prepare
                  Levels 6 through 1 as  shown in  Table 3. All concentrated  stock standard  solutions
                  were made in methanol.

8.   SAMPLE COLLECTION. PRESERVATION AND STORAGE

       8.1     SAMPLE COLLECTION

-------
                                                                           NHSRC/NIOSH SAP

                                                                          Revision Date: 3/30/11
                                                                                 Page 9 of 38
           8.1.1   The exact choice of sampling vessel and procedure is not critical for the analysis and
                  can be adjusted to meet the needs of the situation as long as the different materials
                  have been tested and show no presence  of the target analytes.  As an example for
                  field samples, the field samplers would collect samples with the appropriate wetted
                  wipe (methanol) and place the wipes in a jar with a cap (e.g., 125  mL Nalgene®
                  polypropylene straight-sided jar with a polypropylene screw cap) and ship the jar
                  containing the sample to the laboratory.

           8.1.2   Wipe samples were collected  using Whatman 42 ashless 55 mm circle filter paper.
                  The required analyte spike solution containing the four analytes of interest was added
                  to the surface, allowed to dry, and wiped with each wipe separately. Two wipes were
                  separately wetted with approximately 300 |iL of methanol.  The first wipe is used to
                  wipe the surface in a Z-like pattern horizontally across a defined  surface  (100 cm2)
                  (Figure 18.3).  The second wipe is used to wipe the same surface in a Z-like pattern
                  vertically across a defined surface (100 cm2). Then both wipes are placed  into a 125
                  mL Nalgene polypropylene  straight-sided jar with a polypropylene screw cap.
                  Surrogate (DEA-d8) and methanol solvent (10 mL) are added to the jar.  Because the
                  wipe can lie flat on the bottom of the jar, the solvent fully immerses the wipes. Field
                  and/or matrix blanks are needed, according to conventional sampling practices.

       8.2     SAMPLE STORAGE AND HOLDING TIMES

           8.2.1   Samples should be analyzed as soon as possible.  All samples were refrigerated at 4
                  °C  (±  2 °C) from the time of collection until analysis  unless the samples were
                  analyzed within a 24-hour time period. At the laboratory, samples were stored in the
                  refrigerator at 4 °C (± 2 °C) until requested for analysis.  Samples should be analyzed
                  within 48 hours of collection or as  soon  as possible. Samples from a particular site
                  should be carefully characterized to ensure that there is no interaction with the wipe
                  or specific surface  to cause interferences or degradation of the analytes  after 48
                  hours.  After injection in the LC/MS, the vial septa were replaced and the vials were
                  stored in a refrigerator in case  further analysis was needed. Samples can be stored up
                  to 28 days (Table 2) in the refrigerator at 4 °C (± 2 °C).

9.   QUALITY CONTROL

       9.1     QC requirements include the performance of an initial demonstration of capability (IDC)
               and ongoing QC requirements that must be  met to generate data of acceptable quality
               when preparing and analyzing samples.  This section describes the QC parameters, their
               required frequencies and performance criteria.  A precision and accuracy study (P&A, as
               shown in section 18.2 Attachment) as well as a detection limit study (DL, as shown in
               Table  1 and Section  18.1 Attachments)  must be performed to demonstrate laboratory
               capability  or whenever a major modification is made to  this SAP.   Laboratories are
               encouraged to institute additional QC practices to meet their specific needs.

       9.2     INITIAL DEMONSTRATION OF CAPABILITY (IDC)

               The IDC must be successfully performed prior to the  analysis of field samples.  Prior to
               conducting an IDC, an acceptable Initial Calibration must be generated as outlined in
               Section 10.2.

-------
                                                                  NHSRC/NIOSH SAP

                                                                 Revision Date: 3/30/11
                                                                        Page 10 of 38
    9.2.1   INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND

           Any  time  a new lot  of solvents,  reagents and autosampler vials is used, the
           laboratory  reagent blank (LRB) must  be  demonstrated  to be reasonably free  of
           contamination and that criteria are met in Section 9.4.1. The LRB was used to ensure
           that analytes of interest or other  interferences were not present in the laboratory
           environment, the solvent, or the apparatus.

           NOTE:  Good laboratory practices indicate the  use of a blank during calibration of
           instrumentation to ensure that no carryover occurs between samples.  If the required
           criteria were not met and samples were not free of contamination, then the source of
           the contamination should be identified and eliminated before the performance of any
           analysis.

    9.2.2   INITIAL DEMONSTRATION OF PRECISION AND ACCURACY (P&A)
           NOTE: Because porosity will inevitably have an effect on analyte recovery from the
           surface, accuracy results between calculated values and true values may differ from
           surface to surface.   The precision and accuracy  results are based on Formica®
           (Formica, Cincinnati, OH) surface because the Formica surface has been shown to
           be mostly free of contamination and is a relatively nonporous surface.

           For a precision and accuracy study  (P&A), prepare a check standard containing
           triethanolamine, jV-ethyldiethanolamine, jV-methyldiethanolamine, diethanolamine
           and diethanolamine-d8, near or below the midpoint concentration of the calibration
           range. This check standard should be analyzed with  a minimum of four replicates.
           For this study,  four different concentrations were chosen with seven samples each.
           The check samples were analyzed according to Section 11.

    9.2.3   The average percent recovery (X), standard deviations (a) and the percent relative
           standard deviation (%RSD) of the recoveries were  calculated for each analyte. The
           % RSD value of < 25% should be applied to all analytes.

    9.2.4   MINIMUM REPORTING LEVEL (MRL)

           Establish a target concentration for the MRL based on the intended use of the
           method.  Establish an Initial Calibration (Section 10.2). The lowest CAL standard
           used to establish the initial calibration  must be at or below the MRL concentration. If
           the MRL concentration is too low, ongoing QC requirements may fail repeatedly.
           The MRL is reported in this study as the lowest calibration level.
    9.2.5   Calibration verification (CCV)

           A mid-level sample  from the calibration curve should be analyzed to confirm the
           accuracy of the fit of the calibration curve/standards after the end of sample batches.

9.3     DETECTION AND QUANTITATION LIMITS (DL and QL)
                                      10

-------
                                                                         NHSRC/NIOSH SAP

                                                                       Revision Date: 3/30/11
                                                                               Page 11 of 38

           The procedure for the determination of the laboratory detection and quantitation limits for
           the EPA approach  follows 40 CFR Part  136 Appendix  B  as  described  in  the
           Environmental Response Laboratory Network (ERLN).  Detection  limits (DLs) represent
           the minimum concentration at which there is a high degree of statistical confidence that,
           when the method reports that an analyte is present, that analyte is actually present (i.e., a
           low risk of false positives).   Quantitation limits (QLs) represent the smallest detectable
           concentration of analyte greater than  the detection limit, where the precision and bias
           achieve program objectives.  The DL and QL were determined for each target analyte.
       9.3.1   Determination of laboratory instrument detection limits (IDLs)

               Laboratory instrument detection limits (IDLs) were determined for each instrument
               used for analyses.  Although the determination of the laboratory IDL is not an EPA
               requirement, the laboratory IDL can be used to  establish an estimate  of the initial
               spiking concentration used for determination of the  DL. The laboratory IDL was
               determined for each analyte as a concentration that produced an average signal-to-
               noise (S/N) ratio in the range of 3:1 - 5:1 for at least three replicate injections. For
               example, successively lower concentrations  of the analytes were  injected  until the
               S/N ratio was in the range of 3:1 - 5:1.  Replicates were then injected at that target
               concentration to ensure that the average  S/N of the replicates was within the  3:1 -5:1
               range.

               (NOTE:  S/N ratios must be demonstrated; linearity of S/N ratio with  increasing  or
               decreasing concentration cannot be assumed.)

       9.3.2   Determination of laboratory MDL

               DLs represent the optimal detection achieved by a laboratory in a matrix of interest.
               Formica coupons were used for the determination of the MDL  for  surface  samples.
               The 40 CFR Part 136, Appendix B procedure was followed, particularly with regard
               to  spike levels  used.  Replicate reference matrix samples were  spiked at a level
               between 1-5 times  the estimated detection level (e.g., the IDL, 3 times the standard
               deviation of replicate instrument measurements of the analyte in desired solvent,  or
               the region where there is a break in the slope at the low end of the standard curve).
               The  resulting DL must be  within 10  times  the  spike  level  used, or the DL
               determination would be repeated using a more appropriate spike level.  Full method
               sample preparation procedures to prepare and analyze at least seven replicates of the
               spiked  clean  matrix of interest were used.   Apply  the following equation to the
               analytical results (Student's t-factor is dependent on the number of replicates  used;
               the value 3.14 assumes seven replicates):

MDL = t („-!, l-a = 0.99) X SD

               where

               MDL = method detection limit
                                           11

-------
                                                                   NHSRC/NIOSH SAP

                                                                  Revision Date: 3/30/11
                                                                         Page 12 of 38
           t           = Student's t value for the 99% confidence level with n-1  degrees of
           (n-l,l-a = 0.99)
           freedom (for seven replicate determinations, the Student's t value is 3.143 at a 99%
           confidence level),
           n = number of replicates, and
           SD = standard deviation of replicate analyses.
           a = standard deviation of the percent recovery

          Data for DLs are shown in Table 1 and Attachment  18.1.
9.4    ONGOING QC REQUIREMENTS

    9.4.1   LABORATORY REAGENT BLANK (LRB)

           A reagent blank was prepared and analyzed with each analysis batch, using methanol,
           for confirmation that there were no background contaminants interfering with the
           identification or quantitation of the target analytes. If there was a contaminant within
           the retention time window preventing the determination of the target analyte, the
           source of the contamination should be determined and eliminated before  processing
           samples.

    9.4.2   CONTINUING CALIBRATION CHECK (CCC)

           CCC standards (near the midpoint of the  calibration range) are  analyzed at the
           beginning of each analysis batch, after every twenty field samples, and at the end of
           the analysis batch.

    9.4.3   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM)

           A LFSM is analyzed to  determine that  spike accuracy  for a sample matrix is not
           adversely affected.  If a variety  of sample matrices is analyzed, performance should
           be established for each surface.

        9.4.3.1  Within each analysis batch, a LFSM is prepared and analyzed at a frequency of
                one sample  matrix for every twenty samples.   The LFSM  is prepared by
                spiking a sample with the appropriate amount of analyte AS  (Section 7.2.2).
                Records are maintained of the surface target compound spike analyses, and the
                average   percent recovery (X) and  the standard  deviation  of the percent
                recovery  (a) are calculated.  Analyte recoveries may exhibit  bias  for certain
                matrices. Acceptable recoveries are 50-150% if a low-level concentration near
                or at the  MRL is used. If the accuracy does not fall within this range, check
                with a CCC or prepare a fresh AS solution for analysis.
    9.4.4   SURROGATE STANDARD

           All  samples were spiked with surrogate standard spiking solution as described in
           Section 7.2.1.  An average percent recovery of the  surrogate compound and the
           standard deviation of the percent recovery were calculated and updated regularly.
                                      12

-------
                                                                         NHSRC/NIOSH SAP

                                                                        Revision Date: 3/30/11
                                                                               Page 13 of 38

           9.4.5   MATRIX SPIKE (MS) OR LABORATORY FORTIFIED SAMPLE MATRIX
                  DUPLICATE (LFSMD)

                  Within each analysis batch, a minimum of one MS or LFSMD should be analyzed.
                  Target compound spike  accuracy  in the sample matrix is monitored and updated
                  regularly. Duplicates check the precision associated with sample collection, storage
                  and laboratory procedures. Records are maintained of spiked matrix analyses and the
                  average  percent  recovery  (X) and  corresponding  standard  deviation  (a)  are
                  calculated. MS/LFSMD samples must be  incorporated into the field sampling plan.
                  If the laboratory did not receive MS samples  for determination  of  site-specific
                  precision and accuracy (P&A), the laboratory will evaluate the site data quality based
                  on the Laboratory Fortified Sample Matrix (LFSM) data, if there is sufficient sample
                  in the site samples to conduct an analysis.  MS/LFSMD recovery results will be used
                  for site-specific precision and accuracy (P&A)  data.  LFSM  data were used as
                  MS/LFSMD sample data for this study. RSD values should be < 30% for samples.

           9.4.6   METHOD MODIFICATION QC REQUIREMENTS

                  The analyst may modify the separation technique, LC column, mobile phase
                  composition, LC conditions and MS conditions so as long as all QC and ongoing QC
                  criteria are met. It is the laboratory's responsibility to review the results when
                  method modifications are implemented. If repeated failure occurs, the modification
                  must  be abandoned.
10. INSTRUMENT CALIBRATION AND STANDARDIZATION

   All laboratory equipment should be calibrated according to manufacturer's protocols. Demonstration
   and documentation of acceptable mass spectrometer (MS) tuning and initial calibration is necessary
   prior  to  sample analysis.8  Verification for the tuning of the MS must be  repeated each time
   instrument modification/maintenance is performed and prior to analyte calibration.  After initial
   calibration is successful, a CCC should be performed at the beginning and end of each analysis batch.

       10.1    CALIBRATION OF MAS S SPECTROMETER

              Mass calibration of the mass spectrometer (Waters Acquity™ or equivalent) is performed
              monthly or when mass shifts of more than 0.5 daltons  are noticed by the analyst.  The
              mass calibration file is saved in the mass spectrometer software file folder (MassLynx™
              or similar software). The mass calibration solution used is a mixture of NaCsI provided
              by  the manufacturer.   Other  calibration solutions can also be used per  instrument
              manufacturer's specifications.  The detailed procedure for mass calibration of the mass
              spectrometer  can be  found  in the  software  instruction  manual provided  by the
              manufacturer.

       10.2    INITIAL CALIBRATION FOR ANALYTES

          10.1.1  Optimize  the  [M+H]+  ion for each  analyte by infusing a 500 ng/mL methanol
                  solution directly into the MS at a flow rate of 0.3 mL/min.  The MS  parameters


                                            13

-------
                                                                           NHSRC/NIOSH SAP

                                                                          Revision Date: 3/30/11
                                                                                 Page 14 of 38
                  (voltages, temperatures, gas flows, etc.) are varied until optimal analyte responses are
                  achieved.  Optimize the product ion by following  the  same procedures for the
                  [M+H]+  ion.   Ensure that there  are at least 10 scans across the peak for optimal
                  precision. ESI-MS and MS/MS parameters are presented in Tables 4 and 6.

           10.1.2  Establish LC operating  conditions that  will optimize peak resolution and shape.9
                  Suggested LC conditions (listed in Table  5) may not be optimal for all LC systems.

           10.1.3  The initial calibration contains a seven-point curve using the analyte concentrations
                  prepared in section 7.2.3 and are shown in Table 3.   The lowest calibration curve
                  standard must  be at the MRL.   Depending on the  instrument, sensitivity and
                  calibration curve responses may vary. At a minimum, a  five-point linear or a six-
                  point quadratic calibration curve will be utilized for all analytes.  The coefficient of
                  determination (r2)  of the linear fit should be greater than or equal to  0.98.  The
                  coefficient of determination (r2) of the quadratic curve  should be greater than or equal
                  to 0.99.   A  calibration  curve and an  instrument blank will  be analyzed at the
                  beginning of each batch or daily to ensure instrument stability.10 When quantitated,
                  each calibration point for each analyte should calculate to be within 70-130% of its
                  true value. The lowest CAL standard  should calculate to  be within 50-150% of its
                  true value.  A new curve will  be generated daily. The calibration method is saved
                  and used to quantify all samples.

       10.3    QUANTITATION OF ANALYTES

               The quantitation of the target analytes is accomplished with quantitation  software as it
               relates to each specific instrument (QuanLynx™ or  similar  software).11  An external
               calibration is used along with monitoring diethanolamine-d8 surrogate recovery. Refer to
               Table 4 for the MRM transitions and retention times.
11. ANALYTICAL PROCEDURE

       11.1    SAMPLE PREPARATION

           11.1.1  Samples were collected and stored as described in Section 8.  Surrogate (DEA-d8)
                  and methanol solvent (10 mL) were added to the jar. Sonicate each jar containing the
                  methanol solution for approximately 15 minutes in a water bath at room temperature
                  with no heat required.

           11.1.2  After sonication,  decant the extraction solvent into a  10 cc lock-tip sterile fitted
                  syringe with a Millex® GV syringe driven filter unit, polyvinylidene fluoride (PVDF)
                  filter (0.22 |om), transferring the filtered sample to a sterile 15-mL polypropylene
                  tube (or equivalent).

           11.1.3  Place  the  15-mL polypropylene tube  on the  nitrogen evaporator and set the
                  temperature of the water bath to 50 °C (± 5 °C).

           11.1.4  Concentrate sample in the 15 mL polypropylene tube to < 2 mL using the N-Evap
                  (Thomson Instrument Co., Clear Brook, VA) concentrator.
                                              14

-------
                                                                    NHSRC/NIOSH SAP

                                                                   Revision Date: 3/30/11
                                                                          Page 15 of 38
    11.1.5  Dilute the sample to 2 mL (± 5% accuracy) final volume using methanol. (NOTE:
           More suitable glassware with better accuracy may be used to ensure an exact 2 mL
           final volume.)

    11.1.6  Transfer (via pipette) to a standard 2 mL sample vial.

           NOTE: Calibration standards are not filtered through the syringe-driven filter units
           since no particulates are present.  The filters used in this study were not shown to
           affect analyte concentrations.  If alternate filtering is incorporated, the filters should
           be subjected to QC requirements to ensure they do not introduce interferences  or
           retain the target analytes.

11.2   SAMPLE ANALYSIS/ANALYTICAL SEQUENCE

    11.2.1  Establish Liquid Chromatography/Mass Spectrometry conditions as per guidance
           described in Section 10 and summarized in Tables 4, 5 and 6.

    11.2.2  Prepare a sequence that includes all QC samples and surface  samples.  The first
           sample to be analyzed is a 5 uL injection on column of a blank (methanol).

    11.2.3  The calibration standards, Levels 1  through 7,  are analyzed next. The calibration
           curve and all samples should be analyzed in a low to high concentration regimen  so
           carryover is less of a concern in case the LC cleaning cycle does not clean the system
           adequately between injections. Verify that all analytes have been properly identified
           and quantified using  software programs.  Integrate manually as necessary.  Print
           quantitation reports for the calibration standards.

    11.2.4  Update the calibration  file and print a  calibration report.   Review  the  report for
           calibration outliers and make  area corrections by manual integration, if necessary and
           appropriate.  If corrections have been made, update the calibration file and regenerate
           a calibration report. Alternatively, re-analyze "nonconforming" calibration level(s)
           and repeat the above procedures.

    11.2.5  The first sample analyzed after the calibration curve is a blank to ensure there is no
           carryover.9  If the  initial calibration data are acceptable, begin analyzing samples,
           including QC and blank samples, at their appropriate frequency injecting the same
           size aliquots (5 |iL) under the same conditions used to analyze CAL standards. The
           ending  CCC must have  each  analyte concentration within 30% of the calculated true
           concentration or the affected  analytes from that run must be qualified as estimates or
           the samples must be re-analyzed  with passing criteria to remove the qualification.
    11.2.6  EXCEEDING THE CALIBRATION RANGE:  If the absolute amount of a target
           compound exceeds the working range of the LC/MS system (see Level 7 in Table 3),
           the prepared sample is diluted with methanol and re-analyzed. Care must be taken to
           ensure that there is no carryover of the analyte that has exceeded the calibration
           range. If the amount of analyte exceeds the calibration range, a blank sample should
           be analyzed afterwards to demonstrate no carryover will occur.

    11.2.7  At the conclusion of the data acquisition, use the same software that was used in the


                                       15

-------
                                                                            NHSRC/NIOSH SAP

                                                                          Revision Date: 3/30/11
                                                                                  Page 16 of 38
                  calibration procedure to identify peaks of interest from the retention time windows.
                  Use the  data  software  to examine  the  ion  abundances  of the peaks in  the
                  chromatogram  and  compare   retention  times  with the  retention time  of  the
                  corresponding peak in an analyte standard.

           11.2.8  All qualitative and quantitative measurements are performed as described in Sections
                  9.2 and 9.3. When not being analyzed, samples are stored in the refrigerator at 4 °C
                  (± 2 °C) and protected from light in screw cap top vials equipped with Teflon-lined
                  septa.

12. DATA ANALYSIS AND CALCULATIONS

       12.1    QUALITATIVE AND QUANTITATIVE ANALYSIS

           12.1.1  An external calibration is used  when monitoring  the MRM transitions of each
                  analyte.   Quantitation software (such as QuanLynx™) is  utilized  to  conduct  the
                  quantitation of the target analytes and surrogate standard.  The MRM transitions of
                  each analyte are used for quantitation and confirmation. The MRM transition serves
                  as a confirmation by isolating the precursor ion, fragmenting the precursor ion to the
                  product  ion, and  relating  the  transition to the  retention time in  the calibration
                  standard.11

           12.1.2  Computer  programs  used for  analysis  of  data  include  instrumentation  and
                  quantitation software  (e.g., MassLynx™ with QuanLynx™).  The  manufacturer's
                  quantitation software  manual should be consulted to ensure the proper use of the
                  software. The  quantitation method is  set as an external calibration using the peak
                  areas in ng/mL as long as the analyst is consistent.  Manual integration  may be
                  necessary for some peak areas if the  peak area is not integrated properly (i.e.,  the
                  integration for the peak is not fully performed by the instrument's software, which
                  will be noticeable by visual inspection of each peak).  Inspect all integrated peaks for
                  visible integration errors and manually integrate as necessary to match the integration
                  of other peaks and/or known calibration peaks.  Any manual integration  should be
                  carried  out  by  a qualified  analyst and checked  against quality control procedures
                  (sections 9 and  10.3).

           12.1.3  If the polynomial type is linear and excludes the point of origin, use a fit weighting of
                  1/X in order to give more weighting to the lower concentrations. The retention time
                  window of the MRM transitions must be  within 5% of the  retention  time of the
                  analyte in a mid-range calibration standard.  If this is not true, the calibration curve
                  needs to be re-analyzed to see if there  was a shift in retention  times during  the
                  analysis and the sample needs to be re-injected.  If the retention time  is still incorrect
                  in the  sample,  the analyte is referred  to  as  an unknown.  The coefficient of
                  determination, r2, should be >  0.98  for each  analyte.  If one  of the calibration
                  standards other than the high or low standard causes the curve to be <0.98 this point
                  must be re-injected or a new calibration curve must be analyzed.  If the low and/or
                  high point is excluded, a six-point curve  is acceptable but the calibration  range and
                  reporting limits must be modified to reflect this change.

           12.1.4  If the polynomial type is quadratic, the point of origin is excluded and a fit weighting
                  of 1/X  is used in order to  give more weighting to  the lower concentrations.  The


                                              16

-------
                                                                         NHSRC/NIOSH SAP

                                                                        Revision Date: 3/30/11
                                                                               Page 17 of 38
                 retention time window of the MRM transitions must be within 5% of the retention
                 time of the  analyte in a mid-range calibration  standard.   If this is not true  the
                 calibration curve needs to be re-analyzed to see if there was a shift in retention times
                 during the analysis and the sample needs to be re-injected.  If the retention time is
                 still incorrect in the sample the analyte is referred to as an unknown. The coefficient
                 of determination, r2, should be > 0.99 for each  analyte.  If one  of the calibration
                 standards other than the high or low standard causes the curve to be <0.99, this point
                 must be re-injected or a new calibration curve should be analyzed.  If the low or high
                 point  is excluded, a six-point curve is acceptable using a quadratic fit.  An initial
                 seven-point curve over the calibration range is suggested in the event the low and/or
                 high point must be  excluded to obtain a coefficient of determination > 0.99.  In this
                 event, the calibration range and detection limits must be modified  to  reflect this
                 change.

      12.2   Prior to reporting data, the chromatogram should be reviewed for any incorrect peak
             identification.  If peaks need to be manually adjusted due to incorrect integration by the
             program, clarification of where professional judgment was used to alter the peaks should
             be documented during the data reduction and verification process.

      12.3   All  data packages  will be verified by a qualified  analyst to ensure incorrect peak
             identifications or poor integrations were properly  identified.  The qualified analyst will
             sign off on the narrative and checklist.

13. METHOD PERFORMANCE

      13.1   PRECISION, ACCURACY AND DETECTION LIMITS

          13.1.1  Tables for precision, accuracy and detection limit results for a single laboratory study
                 are presented in Sections 18.1  and 18.2 and Table  1.

      13.2   RECOVERIES AND PRECISION FOR OTHER SURFACE TYPES

          13.2.1  Section  18.2 lists recoveries and precision of target analytes for a variety of other
                 surfaces.

       13.3  PROBLEM ANALYTES AND SURFACES

          13.3.1 TARGET ANALYTES ON PRE-CLEANED AND UNCLEANED SURFACES

                 Target analytes were  spiked on  surfaces and the  wipe  samples were tested  for
                 differences between pre-cleaned surfaces, using methanol, and wipe samples from
                 uncleaned surfaces  (used as received).  When  surfaces  are cleaned prior to analysis,
                 noticeable differences in TEA and DEA recoveries may occur due to a pre-existing
                 presence/contamination of surfaces with these  specific  two compounds (matrix
                 blanks will confirm the presence of TEA and DEA).  Potential matrix effects are also
                 indicated, suggesting laboratories should seek to understand matrix effects occurring
                 in specific samples through  thoughtful choice of MS materials.  Although pre-
                 cleaning surfaces would provide a  more accurate analysis of the true recovery of
                 TEA and DEA, it is not practical in a real scenario.  Analysts  should be aware that
                 these two specific compounds may already be present within the tested sample matrix


                                            17

-------
                                                                           NHSRC/NIOSH SAP

                                                                         Revision Date: 3/30/11
                                                                                 Page 18 of 38
                  and plan accordingly. Wood and painted drywall surfaces resulted in poor recoveries
                  outside  the range of this procedure.  As  a result, the  SAP should not be used to
                  identify these analytes in relation to these specific surfaces. Although porosity of the
                  surface  is most  likely the  culprit  for low recoveries,  further  analysis  should be
                  performed to determine definitive  reasoning of poor recoveries from the surface.
14.  POLLUTION PREVENTION

       14.1   This method utilizes the use of small volumes of organic solvent and small quantities of
              pure analytes, thereby minimizing the potential hazards to both analyst and environment.

       14.2   For  information  about pollution  prevention  that may  be applicable to  laboratory
              operations,  consult "Less  is  Better:  Laboratory  Chemical Management for Waste
              Reduction" available from the American Chemical Society's Department of Government
              Relations and  Science Policy, 1155 16th Street N.W., Washington, D.C., 20036 or on-
              line at http://membership.acs.org/c/ccs/pub_9.htm (accessed November 2009).

 15.  WASTE MANAGEMENT
       15.1   The analytical procedures described in this procedure generate relatively small amounts
              of waste since only small amounts of reagents and solvents are used. Laboratory waste
              management practices must be conducted consistent with all applicable rules and
              regulations, and that laboratories protect the air, water, and land by minimizing and
              controlling all releases from fume hoods and bench  operations. Also, compliance is
              required with any sewage discharge permits and regulations, particularly the hazardous
              waste identification rules and land disposal restrictions.

       15.2   Each laboratory should determine with local officials how to safely dispose of field
              and QC samples. Waste containers should be properly labeled to identify the contents.
              Remember to attach the appropriate chemical waste label and  date the beginning of
              collection before using the container.
16. REFERENCES

       1.  Code of Federal Regulations, 40 CFR Part 136, Appendix B. Definition and Procedure for the
          Determination of the Method Detection Limit - Revision 1.11.

       2.  Federal Advisory Committee on Detection and Quantitation Approaches and Uses in Clean
          Water Act Programs. Final Report. December 28, 2007.

       3.  Lloyd Currie. "Quantitative Determination: Application to Radiochemistry." 1968.
          Anal.Chem 40(3):586-593L.

       4.  Standard Practices for Preparation of Sample  Containers and for Preservation of Organic
          Constituents, American Society for Testing and Materials, Philadelphia. ASTM Annual Book
          of Standards, Part 31, D3694-78.
                                              18

-------
                                                                    NHSRC/NIOSH SAP

                                                                  Revision Date: 3/30/11
                                                                         Page 19 of 38

5.   OSHA Safety and Health Standards, General Industry (29CRF1910). Occupational Safety and
    Health Administration, OSHA 2206 (Revised, July 1, 2001).

6.   Carcinogens-Working with Carcinogens, Publication No. 77-206, Department of Health,
    Education, and Welfare, Public Health Service, Center for Disease Control, National Institute
    of Occupational Safety and Health, Atlanta, Georgia, August 1977.

7.   Safety in Academic Chemistry Laboratories, American Chemical Society Publication,
    Committee on Chemical Safety, 7th Edition. Information on obtaining a copy is available at
    http://membership.acs.org/C/CCS/pub_3.htm (accessed November, 2009). Also available by
    request at OSS@acs.org.

8.   Eichelberger,  J.W.;  Harris,  L.E.;  Budde, W.L. "Reference Compound to  Calibrate  Ion
    Abundance Measurement in Gas  Chromatography-Mass  Spectrometry Systems" Analytical
    Chemistry 1975, 47(7): 995-1000. http://pubs.acs.org/doi/abs/10.1021/ac60357a018.

9.   McNair, N.M.; Bonelli,  E.J. Basic Chromatography;  Consolidated Printing: Berkeley, CA,
    1969; p.  52.

10.  Burke, J.A. "Gas Chromatography for Pesticide Residue Analysis; Some Practical Aspects", J.
    Assoc. Off. Anal. Chem. 1965,48:1037-1058.

11.  Peters, F.T.; Drummer, O.H.; Musshoff, F. "Validation of New  Methods", Forensic Science
    International 2007, 165(2):216-224.
                                       19

-------
17.  TABLES AND VALIDATION DATA
                                                                     NHSRC/NIOSH SAP

                                                                    Revision Date: 3/30/11
                                                                           Page 20 of 3 8

Item
Table 1
Table 2
Table 3
Table 4

Title
Method Parameters
Holding Time Study for Nitrogen Mustard
Degradation Analytes
Concentration of Calibration Standards
MRM Retention Times and MRM Ions and
Number
of Pages
1
1
1
1
Revision
Number
1
1
1
1
Date
Revised
3/2011
3/2011
3/2011
3/2011
             Variable Mass Spectrometer Parameters

Table 5       Gradient Conditions for Liquid
             Chromatography

Table 6       ESf-MS/MS Conditions

Table 7       Materials Tested for the Wipe Analysis of
             Nitrogen Mustard Degradation Products

Table 8       List of Consumable Materials Used During
             Sampling
3/2011


3/2011

3/2011


3/2011
                                          20

-------
                                                                       NHSRC/NIOSH SAP

                                                                      Revision Date: 3/30/11
                                                                            Page 21 of 38
   Table 1.  Method Parameters
FORMICA
Analyte
TEA
EDEA
MDEA
DEA
DL*
ng/fcm2t
0.12
0.06
0.07
0.04
ng/mL
12.32
6.25
6.85
4.37
LOQ
ng/fcm2t
1.23
0.63
0.69
0.44
ng/mL
123.2
62.6
68.5
43.7
MRL
ng/mL
10
10
10
10
*Last DL Study- March 2011.
fng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area
of the coupon (100 cm2).
    Table 2.  Holding Time Sample Stability of Nitrogen Mustard Degradation Analytes
Concentration 50 ng/mL (n =5)

Holding Time
(days)
0
7
14
21
28
TEA
Average
%
Recovery
92.98
93.57
74.93
75.30
78.35
% RSD
6
6
10
7
5
EDEA
Average
%
Recovery
97.56
85.15
87.06
83.84
87.77
% RSD
7
9
6
10
5
MDEA
Average
%
Recovery
94.52
82.36
82.38
82.97
86.60
% RSD
6
8
8
10
5
DEA
Average
%
Recovery
93.47
85.06
82.11
77.84
74.16
% RSD
5
7
6
10
12
DEA-d,
o
Average
%
Recovery
96.32
89.75
85.77
83.51
90.08
% RSD
7
9
11
11
9
                                           21

-------
                                                                      NHSRC/NIOSH SAP
 TableS.  Concentrations of Calibration Standards (ng/tnL)
                                                                    Revision Date: 3/30/11
                                                                           Page 22 of 3 8
Analyte /Surrogate
Triethanolamine
/VEthyldiethanolamine
/VMethyldiethanolamine
Diethanolamine
Diethanolamine-d8
Level
1
10
10
10
10
10
Level
2
25
25
25
25
25
Level
3
50
50
50
50
50
Level
4
100
100
100
100
100
Level
5
250
250
250
250
250
Level
6
350
350
350
350
350
Level
7
500
500
500
500
500
 Table 4.  MRM Ion Transitions, Retention Time and Variable Mass Spectrometer Parameters
Analyte
Triethanolamine
/V-Ethyldiethanolamine
/V-Methyldiethanolamine
Diethanolamine
Diethanolamine-d8 (Surrogate)
Cone
voltage
30
30
30
30
30
MRM mass transition
(parent -product)
150.09 -132.10
134.02 -116.10
120.03 -402.00
106.00 -88.10
114.20 -96.22
Collision
energy (eV)
12
14
12
12
12
Rr
(minutes)
9.7
11.1
13.0
12.1
12.2
Retention times should fall within 5% of the given value; otherwise re-analysis may be necessary.
                                          22

-------
                                                                 NHSRC/NIOSH SAP
                                                                Revision Date: 3/30/11
                                                                      Page 23 of 3 8
Table
5. Gradient Conditions for Liquid Chromatography
Time
(min)
0
1
2
12
16
17
18
21
Flow
(pl/min)
300
300
300
300
300
300
300
300
%
Solution Af
90
90
87
87
85
70
90
90
%
Solution Bft
10
10
13
13
15
30
10
10
     fA: 95% - ACN / 5% - 25mM M^OAC
     f fB: 95% - 25mM NH4OAC (pH 4.22) / 5% - ACN
     Injection volume - 5|iL (recommended)
     * Column Temperature: 30 °C
     *Autosampler Temperature:  15 °C
     * Equilibration time: 3 minutes
     *Column: Atlantis™HILIC silica, 100mm x 2.1mm, 3|im particle size
Table 6. ESI+-MS/MS Conditions
MS Parameter (ESO
Capillary Voltage
Cone Voltage
Extractor
RF Lens
Source Temperature
Desolvation Temperature
Desolvation Gas Flow
Cone Gas Flow
Low Mass Resolution 1
High Mass Resolution 1
Ion Energy 1
Entrance Energy
Collision Energy
Exit Energy
Low Mass Resolution 2
High Mass resolution 2
Ion Energy 2
Multiplier
Gas Cell Pirani Gauge
Inter-Channel Delay
Inter-Scan Delay
Repeats
Span
Dwell
Setting
1.0 kV
See Table 4
2 Volts
0.2 Volts
150 °C
300 °C
800 L/hr
50 L/hr
14.5
14.5
0.5
1
See Table 4
1
15.0
15.0
0.5
-560
3.0 x 10'3 Torr
0.005 seconds
0.005 seconds
1
0.1 Daltons
0.3 Seconds
                                      23

-------
                                                                NHSRC/NIOSH SAP

                                                               Revision Date: 3/30/11
                                                                     Page 24 of 3 8
Table 7. Materials Tested for the Wipe Analysis of Nitrogen Mustard Degradation Products
Material
Glass
Vinyl Tile
Formica
Wood (southern pine, pre-treated)
Galvanized steel
Painted Drywall (BEHR latex paint)
Manufacturer/Vendor
Carolina Glass Co. /Lowe's
Armstrong/Home Depot
Wilsonarf Laminate/Home Depot
Home Depot
McMaster-Carr
BEHR/Home Depot
Table 8. List of Consumable Materials Used During Sampling
Material
Whatman 42 ashless circle filters, 55 mm
125 ml Nalgene polypropylene straight-side jars with
screw caps
10 ml BD safety-lok syringes
Corning 15 ml graduated plastic centrifuge tubes
Millipore 13 mm Millex filter, 0.22 pm PVDF
Waters 1 .8 ml amber glass vials with pre-slit silicone
PTFE screw cap
Vendor
Fisher Scientific (Pittsburgh, PA)
Fisher Scientific (Pittsburgh, PA)
Fisher Scientific (Pittsburgh, PA)
Fisher Scientific (Pittsburgh, PA)
Fisher Scientific (Pittsburgh, PA)
Waters Corp. (Milford, MA)
                                      24

-------
                                                                         NHSRC/NIOSH SAP

                                                                       Revision Date: 3/30/11
                                                                              Page 25 of 3 8
18. ATTACHMENTS
             18.1    Detection and Quantitation Limits

             18.2    Preci sion and Accuracy

             18.3    Illustration depicting the wiping pattern on a 100 cm2 surface
                                            25

-------
                                                                        NHSRC/NIOSH SAP

                                                                       Revision Date: 3/30/11
                                                                              Page 26 of 3 8
    18.1   DL AND QL CALCULATIONS
     PL Calculations for Seven Replicates for Nitrogen Mustard Degradation Analytes
FORMICA

Average Spike
Concentration
(ngyrnL) (n=7)
50.00
Formica Blank
Average Spike
Concentration
(ngyfcm2) (n=7)
0.50
Formica Blank
TEA
Average
Recovery
ng/inL
54.45
26.30
Average
Recovery
(ngyfcm2)
0.54
0.26
%
Recovery
108.9

%
Recovery
108.9

%
RSD
7

%
RSD
7

EDEA
Average
Recovery
ngymL
32.13
0
Average
Recovery
(ngyfcm2)
0.32
0
%
Recovery
64.27

%
Recovery
64.27

%
RSD
6

%
RSD
6

MDEA
Average
Recovery
ngymL
36.92
0
Average
Recovery
(ngyfcm2)
0.37
0
%
Recovery
73.84

%
Recovery
73.84

%
RSD
6

%
RSD
6

fng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area
of the coupon (100 cm2).
FORMICA

Average Spike
Concentration
(ngyhiL) (n=7)
50.00
Formica Blank
Average Spike
Concentration
(ngyfcm2) (n=7)
0.50
Formica Blank
DEA
Average
Recovery
(ng/hiL)
41.05
5.42
Average
Recovery
(ngyfcm2)
0.41
0.05
%
Recovery
82.09

%
Recovery
82.09

%
RSD
3

%
RSD
3

DEA
-------
                                                                        NHSRC/NIOSH SAP

                                                                       Revision Date: 3/30/11
                                                                              Page 27 of 3 8
    18.2    PRECISION AND ACCURACY

       Precision and Accuracy (P&A) data for wipe analysis of nitrogen mustard degradation
       analytes on surfaces, (n = 7 samples at each concentration)
FORMICA

Average Spike
Concentration
(ngAnL) (n=7)
50
75
100
150
Average
Formica Blank
Average Spike
Concentration
(ng/tm2)t
(n=7)
0.50
0.75
1.00
1.50
Average
Formica Blank
TEA
Average
Recovery
(ngAnL)
75.32
71.12
81.11
123.96
23.41
Average
Recovery
(ngytm2)
0.75
0.71
0.81
1.24
0.23
%
Recovery
150.63*
94.83
81.11
82.64
-
%
Recovery
150.63*
94.83
81.11
82.64
-
%
RSD
8
8
6
10
-
%
RSD
8
8
6
10
-
EDEA
Average
Recovery
(ngAnL)
37.11
31.27
65.46
91.24
0
Average
Recovery
(ng/cm2)
0.37
0.32
0.66
0.91
0
%
Recovery
74.23
41.69
65.46
60.82
-
%
Recovery
74.23
41.69
65.46
60.82
-
%
RSD
17
8
4
14
-
%
RSD
17
8
4
14
-
MDEA
Average
Recovery
(ngAnL)
39.49
46.76
66.82
95.14
0
Average
Recovery
(ngytm2)
0.40
0.47
0.67
0.95
0
%
Recovery
78.98
62.34
66.82
63.43
-
%
Recovery
78.98
62.34
66.82
63.43
-
%
RSD
7
11
5
9
-
%
RSD
7
11
5
9
-
*TEA recoveries >150% are consistent with TEA being a native species to this material, as evidenced by
blank coupon samples.  TEA and DEA are not present in sovent blank or wipe blank samples, but are
detected at low levels on the material blank sample, suggesting that TEA and DEA exist for this
material.fng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the
test area of the coupon (100 cm2).
                                            27

-------
                                                                         NHSRC/NIOSH SAP
                                                                        Revision Date: 3/30/11
                                                                               Page 28 of 38
FORMICA

Average Spike
Concentration
(ng/hiL) (n=7)
50
75
100
150
Average
Formica Blank
Average Spike
Concentration
(ng/cm2)t
(n=7)
0.50
0.75
1.00
1.50
Average
Formica Blank
DEA
Average
Recovery
(ngAnL)
56.97
49.13
68.47
99.18
2.41
Average
Recovery
(ngytm2)
0.57
0.49
0.69
0.99
0.02
%
Recovery
113.93
65.51
68.47
66.12
-
%
Recovery
113.93
65.51
68.47
66.12
-
%
RSD
25
6
7
10
-
%
RSD
25
6
7
10
-
DEA-d,
o
Average
Recovery
(ngAnL)
37.49
41.23
62.29
93.05
27.00
Average
Recovery
(ngyfcm2)
0.37
0.41
0.62
0.93
0.27
%
Recovery
74.97
54.97
62.29
62.03
54.01
%
Recovery
74.97
54.97
62.29
62.03
54.01
%
RSD
6
3
5
12
-
%
RSD
6
3
5
12
-
tng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area
of the coupon (100 cm2).
METAL

Average Spike
Concentration
(ng/hiL) (n=7)
50
75
100
150
Average Metal
Blank
TEA
Average
Recovery
(ngAnL)
147.19
145.46
172.36
174.86
117.57
%
Recovery
294.37*
193.95*
172.36*
116.57
-
%
RSD
10
18
13
9
-
EDEA
Average
Recovery
(ngAnL)
42.30
58.43
72.25
91.17
0
%
Recovery
84.60
77.91
72.25
60.78
-
%
RSD
7
8
9
14
-
MDEA
Average
Recovery
(ng/rnL)
42.76
59.01
70.18
71.68
0
%
Recovery
85.52
78.68
70.18
47.79
-
%
RSD
6
6
6
14
-
*TEA recoveries >150% are consistent with TEA being a native species to this material, as evidenced by
blank coupon samples and discussed in section 13.3. TEA and DEA are not present in sovent blank or
wipe blank samples, but are detected at low levels on the material blank sample, suggesting that TEA and
DEA exist for this material.
                                             28

-------
                                                                        NHSRC/NIOSH SAP
                                                                       Revision Date: 3/30/11
                                                                              Page 29 of 3 8
METAL

Average Spike
Concentration
(ngyhiL) (n=7)
50
75
100
150
Average Metal
Blank
DEA
Average
Recovery
(ng/hiL)
64.41
80.65
85.62
97.95
16.28
%
Recovery
128.81
107.54
85.62
65.30
-
%
RSD
5
5
5
4
-
DEA-d8
Average
Recovery
(ngyhiL)
45.32
69.33
72.87
96.35
50.25
%
Recovery
90.64
92.44
72.87
64.23
100.50
%
RSD
10
6
4
3
-
GLASS

Average Spike
Concentration
(ngyhiL) (n=7)
50
75
100
150
Average Glass
Blank
TEA
Average
Recovery
(ng/hiL)
219.24
256.12
256.18
260.51
202.57
% Recovery
438.48*
341.50*
256.18*
173.67*
-
%
RSD
6
10
12
8
-
EDEA
Average
Recovery
(ngyhiL)
25.93
49.72
52.56
78.22
0
%
Recovery
51.87
66.29
52.56
52.14
-
%
RSD
7
5
14
12
-
MDEA
Average
Recovery
(ng/hiL)
30.28
56.65
55.44
83.81
0
%
Recovery
60.56
75.53
55.44
55.87
-
%
RSD
6
4
14
10
-
*TEA recoveries >150% are consistent with TEA being a native species to this material, as evidenced by
blank coupon samples and discussed in section 13.3. TEA and DEA are not present in sovent blank or
wipe blank samples, but are detected at low levels on the material blank sample, suggesting that TEA and
DEA exist for this material.
                                            29

-------
                          NHSRC/NIOSH SAP
                        Revision Date: 3/30/11
                               Page 30 of 38
GLASS

Average Spike
Concentration
(ng/hiL) (n=7)
50
75
100
150
Average Glass
Blank
DEA
Average
Recovery
(ngAnL)
57.51
91.64
77.2
119.25
18.83
%
Recovery
115.01
122.19
77.2
79.5
-
%
RSD
14
4
19
8
-
DEA-d,
o
Average
Recovery
(ngAnL)
29.68
60.67
54.96
91.52
52.47
%
Recovery
59.37
80.9
54.96
61.02
104.94
%
RSD
11
5
21
11
-
VINYL TILE

Average Spike
Concentration
(ngAnL) (n=7)
100
150
200
300
Average Vinyl
Blank
TEA
Average
Recovery
(ngAnL)
29.95
137.26
142.59
224.48
17.6
%
Recovery
29.95
91.51
71.3
74.83
-
%
RSD
25
7
7
10
-
EDEA
Average
Recovery
(ngAnL)
29.34
13.77
52.11
86.98
0
%
Recovery
29.34
9.18
26.06
28.99
-
%
RSD
11
11
8
15
-
MDEA
Average
Recovery
(ngAnL)
29.1
15.95
63.96
75.42
0
%
Recovery
29.1
10.63
31.98
27.79
-
%
RSD
9
7
7
16
-
30

-------
                                                                            NHSRC/NIOSH SAP

                                                                          Revision Date: 3/30/11
                                                                                  Page 31 of 38
VINYL TILE

Average Spike
Concentration
(ngAnL) (n=7)
100
150
200
300
Average Vinyl
Blank
DEA
Average
Recovery
(ngAnL)
36.09
31.91
70.5
78.61
7.00
%
Recovery
36.09
21.27
35.25
26.2
-
%
RSD
13
19
12
10
-
DEA
-------
                                                                            NHSRC/NIOSH SAP

                                                                          Revision Date: 3/30/11
                                                                                  Page 32 of 38
PAINTEDDRYWALL*

Average Spike
Concentration
(ng/hiL) (n=7)
500
Average Drywall
Blank
TEA
Average
Recovery
(ngAnL)
109.63
72.79
%
Recovery
21.93
-
%
RSD
10
-
EDEA
Average
Recovery
(ngAnL)
74.97
0
%
Recovery
14.99
-
%
RSD
17
-
MDEA
Average
Recovery
(ngAnL)
84.91
0
%
Recovery
16.98
-
%
RSD
18
-
PAINTEDDRYWALL*

Average Spike
Concentration
(ngAnL) (n=7)
500
Average
Drywall Blank
DEA
Average
Recovery
(ngAnL)
87.02
23.11
%
Recovery
17.4
-
%
RSD
18
-
DEA-d8
Average
Recovery
(ngAnL)
56.57
39.76
%
Recovery
11.31
79.52
%
RSD
20
-
*Recoveries of all target analytes from this surface are below the acceptable range provided in this SAP.
As a result, the SAP should not be used to identify these analytes in relation to this specific surface.
Although porosity of the surface is most likely the culprit for low recoveries, further analysis should be
performed to determine definitive reasoning of poor recoveries from the surface.
                                              32

-------
18.3  Illustration of wiping pattern on 100 cm  surface
                                                                NHSRC/NIOSH SAP

                                                               Revision Date: 3/30/11
                                                                     Page 33 of 38

in
VI
m
a.
73
OJ

v>
w
ro
Q.
E
co
i

in
VI
03
Q.
*
                                      33

-------
United States
Environmental Protection
Agency
PRESORTED STANDARD
 POSTAGE & FEES PAID
         EPA
   PERMIT NO. G-35
Office of Research and Development (8101R)
Washington, DC 20460

Official Business
Penalty for Private Use
$300

-------