-------
£.
1.8 -
1.6 -
1.4
"oT
= 12
o
0 1 "
"ra
o 0.8-
0.6 -
0.4 -
E. co// 1
^
*>*
•
ly'J
+/
/'
0.2 -]
n I
E. co// 2
/]
J fc
V*
I &
w
I/
I
IH^B
•1
Aldicarb 3
*
i
€'
/'-""^
'
^
Aldicarb 4
— Unitl
i^~^~ — " * Reference
\!
1
I
I
-j—
Unit 2
w
Each section (separated by vertical lines) represents approximately 24 hours.
Figure 6-12. Stage 3 Contaminant Injection Results for Total Chlorine
— Unitl
* Reference
Unit2
Each section (separated by vertical lines; represents approximately 24 hours.
Figure 6-13. Stage 3 Contaminant Injection Results for TOC
37
-------
— Unitl
* Reference
Unit 2
Each section (separated by vertical lines) represents approximately 24 hours.
Figure 6-14. Stage 3 Contaminant Injection Results for Turbidity
— Unitl
* Reference
Unit 2
Each section (separated by vertical lines) represents approximately 24 hours.
Figure 6-15. Stage 3 Contaminant Injection Results for pH
38
-------
.
01
500
480
460-
440-
420 -
400
55
S 380 i
E 360-
340-
320-
300
E CDltl 1
E. coli 2
Aldicarb 3
Aldicart 4
Nor>€TVEco/7ir1ecUon
— Unitl
* Reference
Unit 2
Each section (separated by vertical lines) represents approximately 24 hours.
Figure 6-16. Stage 3 Contaminant Injection Results for Conductivity
The reference method indicated a decrease in pH corresponding to the injection of E. coli and
aldicarb. No change in pH was observed during the Stage 2 injections of aldicarb so the Stage 3
response was unexpected. The WDMP units both detected the change in pH due to the E. coli
injections (pH Event #1 in Figure 6-15), while the change during the aldicarb injections was
detected only by a small inflection of the pH signal (pH Event #2) rather than an obvious peak in
the negative direction, as was observed for the E. coli injections. Finally, conductivity increased
slightly in response to the injection of E. coli. Both the reference method and the WDMP units
detected this change during the first E. coli injection (conductivity Event #1 in Figure 6-16); but,
during the second injection, WDMP Unit 2 generated a zero reading because some air bubbles
had become trapped inside the conductivity sensor (conductivity Event #2). Unit 1 did detect the
change (conductivity Event #3). Note that the second peak (just after the E. coli injection) in the
TOC, turbidity, pH, and conductivity data (labeled non-ETV E. coli injection) reflects a second
injection of E. coli performed that day by the T&E facility staff that was not part of the ETV test.
6.5 Inter-unit Reproducibility
Two WDMP units were compared throughout the verification test to determine whether they
generated results that were similar to one another. This was done using the data collected
whenever a reference sample was collected throughout the verification test. Two evaluations
were performed to make this comparison. First, the results from each sensor from Unit 2 were
graphed on the y-axis, those from Unit 1 were graphed on the x-axis, and a linear regression line
was fitted to the data. For the linear regression analysis, if both units reported the identical result,
the slope of such a regression would be unity, the intercept zero (0), and the coefficient of
determination (r2) 1.0. The slope can indicate whether the results are biased in one direction or
39
-------
the other, while the coefficient of determination provides a measure of the variability of the
results. Second, a t-test assuming equal variances was performed on those same data. The t-test
shows whether the sensors generated statistically similar data. Small p-values (<0.05 at a 5%
confidence level) would suggest that the results from the two units are significantly different
from one another. Table 6-7 gives the slope, intercept, and coefficient of determination for the
inter-unit reproducibility evaluation and the p-value for the t-test performed for each sensor.
Table 6-7. Inter-unit Reproducibility Evaluation
Parameter
Total chlorine
Turbidity
Turbidity (outlier removed)
Temperature
Conductivity
pH
TOC
Slope
0.98
0.77
0.97
0.72
0.92
1.06
0.97
Intercept
0.03
0.09
0.005
7.68
4.19
-0.40
0.31
r2
0.994
0.696
0.881
0.758
0.961
0.919
0.991
t-test p-value
0.779
0.584
0.884
5.5 x 10'6
0.006
0.517
0.374
Shading indicates a significant difference between the two units.
As shown in Table 6-7, the total chlorine, pH, and TOC sensors had coefficients of
determination greater than 0.91 and slopes within 6% of unity, indicating that their results were
very similar and repeatable. Confirming that evaluation, the t-test p-values for those same
parameters were significantly greater than 0.05, indicating that each sensor was generating
statistically similar results. The turbidity measurement, however, generated a slope of 0.77 and a
coefficient of determination of 0.696, suggesting that the results from both units were not well
correlated with one another. However, those results were affected by a single data point in which
Unit 1 generated a result of 3.75 ntu and Unit 2 generated a result of 0.45 ntu. While the reason
for this outlier was not apparent, if this data point were removed, the slope would change to 0.97,
the intercept to 0.005, and the coefficient of determination to 0.881, indicating very similar
results between the two units. In addition, with or without the outlying data point, the t-test
results indicated that the two units were generating statistically similar turbidity data. Figures 6-8
and 6-14 confirm the statistical evaluation of inter-unit reproducibility for turbidity. With the
exception of Unit 1 about halfway through the extended deployment when it drifted high and
displayed a relatively high degree of variability, the results from both units tracked one another
well.
The conductivity meters had a coefficient of determination of 0.961 and a slope of 0.92,
indicating that the data were highly correlated with one another. The t-test generated p-values
significantly less than 0.05, which indicated that the results from the two conductivity sensors
were significantly different. This difference was driven by the small amount of variability in the
conductivity measurements; therefore, the small difference between the means of the two units
was statistically significant. The temperature measurements had a slope of 0.72 and a coefficient
of determination of 0.758, suggesting that the two units were generating statistically different
results. This result did not appear to be driven by outlying temperatures; and the t-test, with a
p-value of less than 0.05, also indicated that the results from each unit were, in fact, different.
Note that the offsets in the conductivity and temperature results (or from any of the parameters)
40
-------
do not affect the performance of the identification algorithm because the baseline is removed and
the identification is performed based only on deviations from baseline.
As discussed for turbidity, the inter-unit reproducibility results for each water quality parameter
were confirmed through a visual evaluation of the figures throughout Chapter 6. As the statistical
results indicated, all the parameters except temperature and conductivity (the two parameters that
had been determined to be significantly different from one another) were nearly overlapping
when plotted on the same axis, indicating that they were, indeed, extremely similar to one
another.
6.6 Contaminant Identification
Thirteen contaminants were injected (in duplicate) during Stage 4 of this verification test.
Section 3.2 describes the straight, single-pass pipe loop that was used. A total volume of 10 L of
each contaminant solution was pumped into the flowing pipe for approximately 20 minutes,
bringing the water to approximately 15 mg/L for each of the contaminants that were injected.
After the leading edge of the injected slug of contaminant reached the WDMP, if the trigger
signal (a proprietary combination of the monitored water quality parameters) exceeded a
specified threshold (trigger event), the EMTS searched the agent library for possible matches.
The EMTS produced an "agent alarm" whenever a trigger event occurred and the deviation in
baseline water quality parameters matched an agent signature in the agent library. When these
signatures were compared with the signatures from the agent library, the quality of the match
was evaluated with a metric called the match angle, which was described in Section 5.4.
Table 6-8 shows the contaminant identification data for each injection that was performed,
including the data from both units tested. A contaminant was never injected without the EMTS
exceeding the trigger threshold and producing a corresponding agent alarm. For both units, the
agent alarms occurred as few as eight times and as many as 79 times during the 20-minute
injection periods. No agent alarms occurred outside of the 20-minute injection periods. As
mentioned previously, each minute-by-minute search of the agent library can result in more than
one agent being identified, which is why more than 20 agent alarms can occur during a
20-minute injection. If the EMTS recognized a deviation from baseline, but the signature did not
match an agent in the library, the trigger event was identified and recorded as an "unknown"
event. Because the leading and trailing edges of the injected contaminant are dynamic, it is
possible that the injection event will generate alarms other than the injected contaminant.
In Table 6-8, the contaminants injected are presented in alphabetical order on left side, and
across the top are the contaminants that were identified by the EMTS agent alarms. There are
more contaminants across the top of the table because three contaminants were identified that
had not been injected. At the time of this test, the EMTS library was populated with 22 contam-
inants. All except one of the 13 contaminants that were injected were among the contaminants in
the EMTS library. The one exception was that pure glyphosate was used for the ETV test, while
the EMTS library was developed using Roundup™, the commercial preparation of glyphosate.
41
-------
Table 6-8. Contaminant Identification—Number and Quality of Matches
Injected
Contaminant
Metha-
midophos
to
W = weak, M = moderate, G = good.
-------
In Table 6-8, for each injected contaminant, the total number of agent alarms is given for each
replicate injection and for each unit. The three columns for each identified agent account for the
quality of the match angle. For example, for the first injection of aldicarb, Unit 1 reported that
nine of the agent alarms were considered weak matches to aldicarb, four were considered
moderate, and there were no good matches. Moving to the right across the table, it can be seen
that aldicarb was identified as colchicine, E. coli, fluoroacetate, and carbaryl by Unit 1 at some
point during the injection. The agent alarms that correctly matched an injected contaminant are
outlined with dark black; and the weak, moderate, and good matches are highlighted with tan,
blue, and yellow, respectively.
The agent alarms resulting from injected contaminants provide an effective way of evaluating
this data. The results for the injection of ferricyanide and lead nitrate clearly were distinct from
the rest of the contaminant injections because, for both units during both injections, agent alarms
were only attributed to those contaminants. The match angles for the ferricyanide alarm fell
entirely in the moderate or weak match quality categories. For the first lead nitrate injection, 39
out of 41 agent alarms were also in the weak or moderate categories, with two in the good match
category; however, for the second lead nitrate injection, 20 out of 43 agent alarms were in the
good match category.
Arsenic trioxide and nicotine were two other contaminants whose data were distinct from the
other contaminants. For these two contaminants, the agent alarms never corresponded to a
correct identification for either unit. Arsenic trioxide was identified most of the time as
glyphosate and less frequently as malathion. Nicotine was most often identified as aldicarb, but
also was identified as colchicine, dichlorvos, and carbaryl. Following ETV test, the Hach
Company updated its agent library with additional data for arsenic trioxide and nicotine to
determine why the results were not as they had anticipated. A summary of their independent
work is given in Appendix A.
The other nine contaminants were sometimes identified correctly and sometimes as another
contaminant. Because of the subjective nature of evaluating the quality of the matches, no
quantitative data analysis that accounts for the match angle will be performed here; but rather a
general discussion of the results presented in Table 6-8. This approach to comparing the agent
alarm results will focus on the data in the context of all the agent alarms that occurred across
both injections and both units to provide an overview of how the EMTS performed.
During the injection of aldicarb, 67 agent alarms were attributed to aldicarb, 45 to carbaryl, and
21 to E. coli. Colchicine, fluoroacetate, malathion, and methamidophos were also identified. Of
the alarms attributed to aldicarb, 21 had good match angles. The rest of the contaminants had
weak or moderate match angles.
Agent alarms during the injection of colchicine were attributed to colchicine and methanol
98 times each and to dichlorvos 78 times. Of the colchicine alarms, 71 were good matches, while
for the dichlorvos alarms, 64 were good matches. For the methanol alarms, 27 fell into that
category.
For the dicamba injection, 51 agent alarms were correctly attributed to dicamba compared with
10 or fewer agent alarms attributed to colchicine, dichlorvos, mercuric chloride, and methanol.
Fifteen of the 51 dicamba agent alarms were good matches.
43
-------
During the dichlorvos injection, 21 agent alarms correctly identified the contaminant, while
78 identified the contaminant as colchicine and 90 as methanol. All but one of the dichlorvos
alarms indicated weak matches, while several alarms for colchicine and methanol indicated good
matches.
The E. coli injection was correctly identified 20 times with weak or moderate match angles,
while it was identified as malathion 53 times, including seven good match angles. Also
identified during the E. coli injection were arsenic trioxide, colchicine, glyphosate, methanol,
carbonyl, and methamidophos between two and 11 times, mostly with weak and moderate match
angles.
The fluoroacetate injection generated 49 correct agent alarms, with all but two in the weak or
moderate match categories. Fluoroacetate was much less frequently (six or fewer times each)
identified as colchicine, methanol, malathion, and methamidophos.
The injection of glyphosate generated seven correct agent alarms even though Roundup™, the
commercial preparation of glyphosate that also includes other organic chemicals, rather than
pure glyphosate, was included in the EMTS library. Because of this, Roundup™ has a different
water quality parameter signature than pure glyphosate. This injection was also identified as
aldicarb (one time), dicamba (17 times), mercuric chloride (30 times), malathion (20 times), and
methamidophos (11 times). The only alarms that were good matches were for methamidophos.
After the ETV test, Hach updated its agent library with data for pure glyphosate. A summary of
Hach's independent work is given in Appendix A.
The mercuric chloride injection produced an almost equal number of agent alarms identifying it
correctly and as dicamba. Eighty-seven agent alarms were for mercuric chloride, with 16 good
matches and 56 moderate matches; and 86 alarms were attributable to dicamba, with 16 good
matches and 57 moderate matches.
Methanol was correctly identified 26 times, with 14 good matches. However, it was identified as
colchicine 32 times with five good matches and as dichlorvos 25 times with six good matches.
To summarize, the EMTS accuracy for detecting an injected contaminant was 100%; that is, in
all cases, the injection of a contaminant caused a deviation from the baseline measurement of the
water quality parameters significant enough to cause a trigger event resulting in an agent library
search. For 11 out of 13 contaminants, at some time during the injection, the correct contaminant
was identified during the search of the EMTS library. Two method blank injections of pipe loop
water did not cause trigger events; and, therefore, no injection was detected.
The data in Table 6-8 are difficult to interpret, but give a complete report of EMTS performance,
including total number of agent alarms, agent alarms that were attributable to the injected
contaminant (correct identification), and those that were not. Table 6-8 also shows the quality of
match that the agent alarms represented. To provide a more concise way of presenting the data,
the fraction of agent alarms attributable to the injected contaminant (correct identification) was
calculated for each injection. This fraction was called the classification rate and is defined in
Section 5.4. (Note that this approach does not take into consideration the match angle of each
agent alarm.) To present the data succinctly, the classification rates were divided into five levels,
which are shown in Table 6-9. Level 5 represents a classification rate of greater than 70%, Level
44
-------
4 between 31% and 69%, Level 3 between 1% and 30%, Level 2 indicates that the injected
contaminant was not correctly identified but other contaminants were identified, and Level 1 was
reserved for instances when no injections were detected. During this verification test, the Level 1
conditions were met only when two method blanks of pipe loop water were injected.
Contaminants with mostly Level 4 and 5 classification rates included dicamba, ferricyanide,
fluoroacetate, and lead nitrate. Those with mostly Level 4 classification rates included aldicarb,
colchicine, mercuric chloride, and methanol. Glyphosate, dichlorvos, and E. coli each had a
mixture of Level 2, 3, and 4 classification rates. Arsenic trioxide and nicotine had classification
rates of 2, which indicates that those two contaminants were not correctly identified during their
injection. As mentioned above, see the appendix to this report for additional data on glyphosate,
arsenic trioxide, and nicotine.
Table 6-9. Classification Rate Levels
Injected
Contaminant
Aldicarb
Arsenic trioxide
Colchicine
Dicamba
Dichlorvos
E. coli
Ferricyanide
Fluoroacetate
Glyphosate
Lead nitrate
Mercuric chloride
Methanol
Nicotine
Unit
4
2
4
4
4
3
5
5
4
5
4
4
2
Injection 1
1 Unit 2
4
2
4
5
3
2
5
5
3
5
4
4
2
Injection
Unitl
4
2
4
5
3
4
5
4
2
5
4
4
2
2
Unit 2
4
2
4
5
2
2
5
4
2
5
4
3
2
Level 5 = >70% correctly identified
Level 4 = 31-70% correctly identified
Level 3 = 1-30% correctly identified
Level 2 = 0%, other contaminants identified
Level 1 = 0%, no contaminant identified
Evaluating the differences between the performance of individual EMTS units in accurately
identifying injected contaminants is difficult because any differences between the two units are a
result of the monitoring data that is input to the algorithm. Repeatability of that data was
discussed previously. Presumably, because this is a software application, if identical data were
input, identical results would be generated.
6.7 Ease of Use and Data Acquisition
Hach Company staff performed all maintenance on the WDMP and EMTS units. They recorded
any maintenance activity they performed on either of the units in a logbook. The WDMPs did
45
-------
not require daily operator attention. Throughout the verification test, Hach Company staff
periodically adjusted the flows on the turbidity and total chlorine meters as needed to keep them
at the required levels and rebooted the EMTS when the real-time display was not displaying data
properly.
Reinitialization (i.e., rebooting the EMTS) occurred almost daily for Unit 2 for the first week or
so of Stage 3, but thereafter, it only was necessary one or two times. This was required when the
real-time display was not functioning properly. The sample cuvettes within the chlorine monitors
were cleaned four times throughout the verification test (twice during extended deployment) to
maintain accurate measurement. This process took approximately 15 minutes. The TOC
analyzers were calibrated three times throughout the test, the reagents were changed out once,
and the TOC manifold was cleaned two times: once after nitrogen flow had actually been
blocked and once after the nitrogen supply had run out. According to the maintenance records,
Hach Company staff cleaned the turbidimeter lines and checked its calibration two times
throughout the verification test. The conductivity data from one contaminant injection was lost
because of an air bubble. This was remedied by opening the conductance meter to release the
bubble.
The data were downloaded from the EMTS using a USB port. The data were in a comma-
delimited format that was easily opened into a spreadsheet. Overall, some of the regular
maintenance such as cleaning the chlorine meter cuvette and turbidimeter and calibrating the
TOC analyzer would have to be performed regularly if this system was placed in a remote
location, requiring periodic site visits.
46
-------
Chapter 7
Performance Summary
Evaluation Parameter
Stage 1—
Accuracy
Stage 2—
Response to
Injected
Contaminants
Stage 3 —
Accuracy During
Extended
Deployment
Stage 3—
Accuracy After
Extended
Deployment
Stage 3—
Response to
Injected
Contaminants
Injection
Summary
Inter-unit
Reproducibility
(Unit 2 vs. Unit 1)
Units 1 and 2, range
of %D (median)
Nicotine
Arsenic
trioxide
Aldicarb
Reference
WDMP
Reference
WDMP
Reference
WDMP
Units 1 and 2, range
of %D (median)
Unit 1, %D
Unit 2, %D
E. coli
Aldicarb
Reference
WDMP
Reference
WDMP
Total
Chlorine
-47.4 to 4.5
(-3.9)
-
-
-
-
-
-
-15.9 to 6.9
(-3.2)
-4.9
-4.9
-
-
-
-
Turbidity
-53.9 to -1.3
(-34.1)
(a)
+
(a)
+
(a)
+
-81.1 to
245.5 (-21.3)
-5.9
-11.8
_i_(b)
+
+(b)
+
Tem-
perature
-3.0 to 44.3
(-0.2)
NC
NC
NC
NC
NC
NC
-7.4 to 8.5
(-0.1)
-0.2
4.6
NC
NC
NC
NC
Conductivity
-15.5 to 8.1
(2.2)
NC
NC
+
+
NC
NC
-1.8 to 9.6
(4.8)
6.7
0.3
+
+
NC
NC
pH
-6.6 to 3.1
(0.9)
NC
NC
+
+
NC
NC
-2.7 to 0.5
(-0.9)
-2.2
0.2
-
-
-
-
TOC
-64.7 to 147.5
(-14.8)
+
+
NC
NC
+
+
-47.3 to 103.0
(-6.9)
-20.5
3.4
+
+
+
+
Total chlorine and TOC were dramatically affected by injections of nicotine, E. coli, and aldicarb; and
turbidity, pH, and conductivity were affected by some or all of the injections, but not as consistently as
total chlorine and TOC. Aldicarb altered the pH during Stage 3, but not Stage 2.
Slope (intercept)
r2
p-value
0.98
(0.03)
0.994
0.779
0.97
(0.005)(c)
0.88 l(c)
0.884(c)
0.72
(7.68)
0.758
5.5 x 10'6
0.92
(4.19)
0.961
0.006
1.06
(-0.40)
0.919
0.517
0.97
(0.31)
0.991
0.374
With the exception of temperature and conductivity, both units generated similar results.
Stage 4—
Contaminant
Identification
Each time a contaminant was injected, the EMTS detected a deviation in baseline conditions, causing a
"trigger event." Eleven of 13 contaminants were correctly identified at some point during the injection
time. Ferricyanide and lead nitrate were identified correctly 100% of the time. The rest of the injected
contaminants were identified as a contaminant other than themselves at some point throughout the
duration of the injection. Only nicotine and arsenic trioxide were never correctly identified.
Ease of Use and
Data Acquisition
Neither the WDMPs nor the EMTSs required daily operator attention. Hach Company staff adjusted the
flows on the turbidity and total chlorine meters as needed to keep them at the required levels and
rebooted the EMTS when it was not displaying data properly. The chlorine sensors and turbidimeters
needed periodic cleaning, and the TOC analyzer was calibrated three times.
-------
Chapter 8
References
1. Test/QA Plan for Verification of Multi-Parameter Water Monitors for Distribution Systems,
Battelle, Columbus, Ohio, August 2004.
2. U.S. EPA, EPA Method 150.1, pH, in Methods for Chemical Analysis of Water and Wastes,
EPA/600/4-79/020, March 1983.
3. American Public Health Association, et al, SM 2510, Conductivity, in Standard Methods for
the Examination of Water and Wastewater. 19th Edition, Washington, D.C., 1997.
4. American Public Health Association, et al., SM 4500-C1-G, Total Chlorine, in Standard
Methods for the Examination of Water and Wastewater, April 13, 2004.
5. U.S. EPA, EPA Method 415.1, Total Organic Carbon, in Methods for Chemical Analysis of
Water and Wastes, EPA/600/4-79/020, March 1983.
6. U.S. EPA, EPAMethod 170.1, Temperature, inMethodsfor Chemical Analysis of Water and
Wastes, EPA/600/4-79/020, March 1983.
7. U.S. EPA, EPA Method 180.1, Turbidity, in Methods for Chemical Analysis of Water and
Wastes, EPA/600/4-79/020, March 1983.
8. Quality Management Plan (QMP)for the ETV Advanced Monitoring Systems Center,
Version 5.0, U.S. EPA Environmental Technology Verification Program, Battelle, Columbus,
Ohio, March 2004.
48
-------
Appendix A
Hach Company Review
The following text summarizes the results acquired by the Hach Company after review of the
ETV test results. This work was performed at its facility without EPA or Battelle QA oversight.
The results should not be considered part of the ETV testing. Questions about these results
should be directed to representatives of Hach Company.
To test whether the EMTS would identify pure glyphosate if it had been populated with data
attained during the injection of that chemical, the water quality parameter fingerprint of pure
glyphosate was added to the EMTS agent library; the contaminant previously called glyphosate
was renamed Roundup™; and the identification algorithm was reapplied to the original ETV
data. Table A-l shows that, during the original ETV evaluation, very few of the agent alarms
during the injection of glyphosate were reporting glyphosate; and, if they did, the match angles
were typically weak. After Hach's update, the vast majority of alarms were reported as
glyphosate, with mostly strong match angles.
In response to the results of the ETV test, The Hach Company updated the EMTS agent library
by including additional arsenic trioxide data and then reanalyzed the original ETV test data.
Table A-l shows that, during the ETV evaluation, arsenic trioxide was never identified during its
injection, while glyphosate (renamed Roundup™ for this reanalysis) was identified frequently.
The updated data from Hach show that some agent alarms were reported as arsenic trioxide, but
typically with low match angles. Also, Roundup™ was still identified frequently. The difficulty
in identifying arsenic trioxide may be due to its partial solubility in water, making it difficult to
maintain a consistent level during the injections into a pipe. The Hach Company noted that it
also had difficulty in maintaining a consistent suspension for its agent library development.
The Hach Company previously (during non-ETV testing) identified nicotine with the EMTS
rather successfully. It was noted that one difference between the nicotine solution used during
the ETV test and that used during the agent library development was how vigorous the stirring
had been during solution preparation. For ETV testing, the solution was stirred with a stirring
attachment on an electric drill as opposed to a small stir bar used by the Hach Company during
library development. The Hach Company performed an experiment both with and without
vigorous mixing and determined that the vigorous mixing caused the basic form of nicotine to
react with atmospheric carbon dioxide to form the neutralized form of nicotine, which had a
lower characteristic pH than the basic form of nicotine. The EMTS agent library signature was
updated to include the vigorously mixed nicotine, and the ETV data was reanalyzed as pre-
viously described for glyphosate and arsenic trioxide. The Hach Company did not provide the
raw data for these results, but they did indicate that nicotine was identified with strong match
angles during their independent testing.
A-l
-------
Table A-l. Comparison of ETV Data and Data Updated by Hach
Injected
Contaminant
Inj.
#
Unit
Total*
of IDs
Quality of angle match
Glyphosate
(ETV results)
Glyphosate
(Hach Update)
Arsenic
Trioxide (ETV
results)
Arsenic
Trioxide (Hach
Update)
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
9
41
13
24
40
44
60
59
19
13
10
18
24
20
25
18
Aldicarb
W
0
M
1
0
G
0
3
Arsenic
Trioxide
W
0
3
1
1
5
0
4
6
0
1
M
3
2
0
1
0
2
0
1
3
G
2
1
0
0
0
0
0
0
0
0
1
0
Dicamba
W
0
5
2
8
2
5
8
M
0
1
0
0
0
1
0
G
1
0
0
0
0
0
0
Glyphosate
W
1
4
0
0
3
4
4
2
16
9
6
2
1
M
1
0
0
4
3
7
7
3
3
3
6
5
G
1
0
0
17
20
9
20
0
0
1
4
0
Mercuric
Chloride
W
3
9
9
9
3
9
M
4
0
5
0
4
5
G
0
0
0
0
0
0
Malathion
W
4
3
2
1
4
2
2
1
1
6
1
M
0
10
0
1
0
0
9
1
0
0
0
G
0
0
0
0
0
0
0
0
0
0
0
Metha-
midiphos
W
1
2
1
2
M
0
1
0
2
G
0
7
0
6
Roundup
W
M
G
NA
0
2
2
0
1
0
NA
5
8
5
3
7
6
6
6
6
0
5
5
>
NA = Contaminant not injected during the ETV test.
-------