-------
6.3.5 E. coli
Table 6-3h presents the false negative and false positive results for E. coli. The number of
positive samples out of the total replicates analyzed is presented in the table. No false positive or
false negative samples were found in any of the sample matrices.
Table 6-3h. E. coli False Positive/Negative Results
Sample Type
Contaminant-Only PT
Samples
Interferent PT
Samples
DW Samples
False Positive Rate
False Negative Rate
Sample
DI water
DI water
DI water
DI water
0.5 mg/L humic acid and
0.5 mg/L fulvic acid
0.5 mg/L humic acid and
0.5 mg/L fulvic acid
2.5 mg/L humic acid and
2.5 mg/L fulvic acid
2.5 mg/L humic acid and
2.5 mg/L fulvic acid
OHDW
OHDW
CADW
CADW
FLOW
FLDW
NYDW
NYDW
Concentration'*0
(cfu/mL)
2xl03
5xl03
IxlO4
5xl04
Blank
IxlO4
Blank
IxlO4
Blank
IxlO4
Blank
IxlO4
Blank
IxlO4
Blank
IxlO4
Positive Results Out of
Total Replicates
4/4
4/4
4/4
4/4
0/4
16/1600
0/4
4/4
0/4
4/4
0/4
4/4
0/4
4/4
0/4
4/4
0/24
0/52(b'c)
(a) Sample solutions were prepared at 2, 5, 10, and 50 times the vendor-stated system LOD from stock solutions based
on the enumeration data (see Table 4-1).
(b) One set of spiked 0.5 mg/L humic and fulvic acid replicates were suspected of having cross-contamination problems.
The samples were not rerun, so the results are not reported here for consistency with other rerun reporting.
(c) The infective/lethal dose for E. coli was below the system LOD and thus not included in this calculation.
39
-------
6.4 Precision
The performance of the F. tularensis Target 1 assay within sample sets of four replicates was
consistent. Only one set of replicates, that for interferent samples spiked with 2.5 mg/L humic
and fulvic acids, was inconsistent, with inconclusive results for three replicates and a negative
result for one replicate. All other samples showed the same results within a set of replicates.
Thus, for F. tularensis Target 1, one of the 21 sets of replicates that were analyzed was
determined to be inconsistent, indicating that 95% of the sample sets showed consistent results
among the replicates.
The precision of the F. tularensis Target 2 for the R.A.P.I.D.® System was not as consistent as
that for Target 1. Results were inconsistent within three sets of replicate samples for Target 2. In
the unspiked CA DW sample, the results for one replicate were inconclusive while the other three
were negative. The same was true for spiked CA DW. In one set of spiked 0.5 mg/L each humic
and fulvic acids, one negative result was found amongst three positive responses. All other
samples showed the same results within a set of four replicates. Thus, for F. tularensis Target 2,
three of the 21 sets of replicates that were analyzed were determined to be inconsistent, indicating
that 86% of the sample sets showed consistent results among the replicates.
The performance of the R. A.P.I.D.® System within sample sets of four replicates for Y. pestis
Targets 1 and 2 was very consistent. For all 21 sets of replicate samples for both targets, all
replicates showed the same results within the sample set. Thus, for Y. pestis Targets 1 and 2,
100% of the sample sets showed consistent results among the replicates.
As with F. tularensis, the consistency of the sample sets for B. anthracis were mixed for the three
targets that were evaluated. For B. anthracis Target 1, five sample sets had inconsistent results.
These samples included two PT samples (2 x 103 cfu/mL and the infective/lethal dose), the spiked
CA DW, the spiked FL DW, and the spiked NY DW samples. For the infective/lethal dose of
B. anthracis, three of the four samples had inconclusive results, while one sample was negative.
The infective dose of B. anthracis was below the LOD for the R. A.P.I.D.® System for this
bacteria, so the discrepancy between replicate samples likely has more to do with this fact than
the actual precision of the system. At 2 x 103 cfu/mL, three positive results and one inconclusive
result were observed. The responses for spiked CA DW were mixed, with two positive replicates,
one negative, and one inconclusive replicate response. Spiked FL DW had three positive results
and one inconclusive result, while spiked NY DW had three positive and one negative response
for B. anthracis. These five inconsistent replicate sets indicate that 76% (16/21) of the sample
sets analyzed for B. anthracis Target 1 showed consistent results among the replicates.
The number of consistent replicate sample sets for B. anthracis Target 2 was higher than that for
Target 1. Three of the 21 replicate sets of samples showed inconsistent results for Target 2. As
with Target 1, the infective/lethal dose of B. anthracis showed varied results, with two incon-
clusive and two negative responses. Again, the infective dose of B. anthracis was below the LOD
for the R. A.P.I.D.® System for this bacteria, so the discrepancy between replicate samples likely
has more to do with this fact than the actual precision of the system. Replicates at 2 x 103 cfu/mL
again showed inconsistent results, with three positive and one negative response. The remaining
inconsistent sample set was spiked CA DW. For these replicates, one was positive, one was
40
-------
negative, and the remaining two were inconclusive. Overall, 86% (18/21) of the sample sets
analyzed for B. anthracis Target 2 showed consistent results among the replicates.
The consistency within sample sets for the B. anthracis assays was the highest for Target 3. As
with Targets 1 and 2, the results for the infective/lethal dose of anthrax were inconsistent, with
two negative and two inconclusive results. The remaining 20 of the 21 sets of replicate samples
showed consistent results among the replicates. All four spiked CA DW samples were
inconclusive. Thus, for B. anthracis Target 3, one of the 21 sets of replicates that were analyzed
was determined to be inconsistent, indicating that 95% (20/21) of the sample sets showed
consistent results among the replicates.
For Brucella suis, two out of 21 sample sets were inconsistent. The replicates at 2 x 103 cfu/mL
showed two positive and two inconclusive responses. Three of the replicates at the infective/
lethal dose were negative, while one was inconclusive, resulting in an inconsistent set of
replicates. As with B. anthracis, the infective/lethal dose of Brucella suis is below the vendor-
provided system LOD for this bacteria. All other sample sets showed consistent results within a
set of replicates. Thus, 90% (19/21) of the sample sets showed consistent results among the
replicates.
Only 20 sets of replicate samples were considered for the precision calculations for E. coli (see
Table 6-3h). For all 20 sets of replicate samples, all replicates showed the same results within the
sample set. Thus, for E. coli, 100% of the sample sets showed consistent results among the
replicates.
6.5 Interferences
6.5.1 Interferent PT Samples
In both the 0.5 mg/L and 2.5 mg/L humic and fulvic acid solutions, both spiked with the bacteria
of interest and unspiked, the R. A.P.I.D.® System provided expected results for Y. pestis Targets 1
and 2; B. anthracis Targets 1, 2, and 3; Brucella suis; and E. coli. In the absence of the bacteria,
all samples for these bacteria (and all associated targets) tested negative; in the presence of the
bacteria, all samples tested positive. In the case of F. tularensis Target 1, all unspiked 0.5 mg/L
humic and fulvic acid samples returned negative results, while all spiked 0.5 and 2.5 mg/L humic
and fulvic acid samples returned positive results. The unspiked 2.5 mg/L humic and fulvic acid
replicates returned one negative response and three inconclusive results, where the split samples
of each of these three replicates had a positive and a negative response. All positive samples had
low amplification and late crossing points. The same sample DNA, when analyzed on the
R.A.P.I.D.® 7200 instrument using F. tularensis Target 2, returned all negative results. The
discrepancy between the targets could be related to the different sensitivities of the two targets
and/or potential cross-contamination. The initial spiked and unspiked 0.5 mg/L and 2.5 mg/L
humic and fulvic acid samples tested using F. tularensis Target 2 returned all positive results
when the sample was spiked and all negative results when the sample was unspiked.
Discrepancies in the results for the additional 0.5 mg/L spiked samples that were tested are
discussed below.
41
-------
As discussed in Section 3.2.1, four solutions of humic and fulvic acids at 0.5 mg/L each spiked
with each contaminant at lx!04cfu/mL were prepared in addition to the initial 0.5 mg/L and
2.5 mg/L humic and fulvic acid solutions. Each solution was put through the DNA extraction and
purification procedure, and then four replicates from each of the four purified DNA solutions
were analyzed using the R. A.P.I.D.® System. These samples were included in the verification test
in an effort to evaluate the efficacy of the DNA extraction and purification procedure in the
presence of inhibitory substances. These samples also contribute to the precision evaluations of
the R.A.P.I.D.® System. For F. tularensis Target 1; Y. pestis Targets 1 and 2; B. anthracis
Targets 1, 2, and 3; Brucella suis; and E. coli, all of the samples tested resulted in positive
responses. Thus, 20 out of the 20 (16/16 for E. coli) spiked 0.5 mg/L humic and fulvic acid
samples tested resulted in positive responses for each bacteria. For F. tularensis Target 2, one of
the 20 replicates tested produced a negative instead of positive reponse.
6.5.2 Drinking Water Samples
The R. A.P.I.D.® System DW sample results for F. tularensis, Y. pestis, B. anthracis,
Brucella suis, and E. coli are presented in Tables 6-3a through 6-3h. In general, the R.A.P.I.D.®
System showed positive results for each set of replicates for the spiked DW samples and negative
results for each set of replicates for the unspiked samples, with a few exceptions. B. anthracis
Targets 1, 2, and 3 as well as F. tularensis Targets 1 and 2 generated mixed results in spiked CA
DW. B. anthracis Target 1 also showed inconclusive and negative results in spiked FL and NY
DW, respectively. For the detection of F. tularensis Target 2 in unspiked CA DW, one of the four
replicates had an inconclusive result. Analysis of the CA DW did not indicate the presence of
F. tularensis. The possibility of cross-contamination causing the inconclusive results for
unspiked CA DW sample cannot be ruled out.
The contaminant-only PT samples at IxlO4 cfu/mL, the level at which the DW samples were
spiked, showed consistent positive responses across all bacteria. The interferent PT samples at
both 0.5 mg/L and 2.5 mg/L humic and fulvic acids also spiked at IxlO4 cfu/mL showed
consistent positive responses for all replicates across all bacteria, except for four replicates total
for F. tularensis Targets 1 and 2. The consistency of responses in these PT samples, as well as
other contaminant-only PT samples above the system LOD, would seem to indicate that some of
the DW matrices used in this test, in particular CA DW, may have inhibitory or other
confounding effects on the PCR process for the R. A.P.I.D.® System.
6.6 Other Performance Factors
The R. A.P.I.D.® System was operated by the same Battelle technician throughout the verification
test. This technician had prior PCR experience and was trained by Idaho Technology over the
course of one day in the operation of the R. A.P.I.D.® System before testing began. This training
included the use of the ITI1-2-3 Flow Kit, the R.A.P.I.D.®7200 instrument, and the freeze-dried
reagents. The Battelle technician was familiar with general DNA extraction and purification
techniques, PCR reagent preparation techniques, and general thermal cycler operation, as well as
general PCR theory, prior to training. The overall operation of the R. A.P.I.D.® System was
straightforward, and the experienced technician found the system easy to use and had no major
42
-------
difficulties operating the R. A.P.I.D.® System. The R. A.P.I.D.® System is designed to be used by
operators without prior experience who have only gone through training conducted by Idaho
Technology. Some features of the R. A.P.I.D.® System software are designed to complement such
an inexperienced user, though the Advanced Options were used in this test.
The freeze-dried reagents provided for easy PCR setup because all of the components needed to
fill two capillaries were contained within one vial. The "unknown" reagents and positive and
negative control reagent vials were packaged together and color coded. It was important to
throughly mix the reconstituted reagents before pipetting them into the capillaries. Prior
experience using pipettes would likely make the PCR setup easier and reduce potential pipetting
errors. The DNA extraction procedure using the ITI1-2-3 Flow Kit was straightforward and easy
to follow, though there were a fair number steps involved in the process. The transfer pipettes
provided with the ITI 1-2-3 Flow Kit were more rudimentary than typical laboratory pipettes, but
they worked well because exact amounts of buffer were not needed for the extraction procedure.
The glass capillaries used to hold the sample were easy to work with but posed some problems.
At one point, it became difficult to put the capillaries into the slots (rotors) in the carousel, and
the technician broke multiple capillaries trying to load the instrument. Capillaries also broke
during two PCR runs. After instruction from Idaho Technology, the rotor tool (provided with the
system) was used to clean out the capillary slots in the carousel. After the cleaning, the capillaries
posed no further problems. The R.A.P.I.D.® System software was user friendly and easy to use
with clearly labeled buttons. The technician easily learned to use the Advanced Options feature,
which was used to evaluate the results instead of the user interface intended for inexperienced
operators. In Advanced Options, the software returned crossing points for all samples that had
crossed the background fluorescence, but it was up to the technician to verify that the sample was
positive. An understanding of PCR theory and amplification would make such interpretations
easier.
All testing was performed in a laboratory setting because the R.A.P.I.D.® System as used in this
verification test is not field portable. The R. A.P.I.D.® System is intended to be field portable, and
in fact the instrument itself as used for this test could be easily transported to the field. However,
to perform the DNA extraction and purification of 5 mL of water, a 50 mL capacity centrifuge
had to be used. For this reason, the R. A.P.I.D.® System as used in this test was not considered
field portable. Three different testing areas were required in each laboratory to operate the
R. A.P.I.D.® System: one for the DNA purification, one for the reagent preparation, and one for
the instrument operation. All of the reagents, including those for the ITI 1-2-3 Flow Kit and the
freeze-dried reagents, were stable at room temperature. The reagents came packaged in
individual, vacuum-sealed foil pouches containing 20 vials (four positive controls, two negative
controls, and 14 unknown vials) along with sterile PCR grade water. The R.A.P.I.D.® 7200
instrument was approximately 20 inches x 14 inches x 10 inches its own hardened case, with a
backpack for carrying. It came with its own laptop, which can be placed in the backpack.
R tularensis and E. coli samples were tested in a BSL-2 laboratory; while Y. pestis, B. anthracis,
and Brucella suis were tested in a BSL-3 laboratory. Because live bacteria were being handled,
special safety requirements and protocols had to be implemented in both the BSL-2 and BSL-3
laboratories. Some of these requirements impacted the analysis time for the R. A.P.I.D.® System
and are inherently present in any throughput estimations for this verification test. Thus, such
43
-------
performance factors mentioned here also incorporate the safety and facility requirements
necessary for this test.
A total of 92 or more samples (including method blanks) were tested for each bacteria using the
R. A.P.I.D.® System. On average, the DNA extraction and isolation step for between three and
nine solutions took approximately 2 hours. The PCR setup steps, including reconstituting the
freeze-dried reagents and loading the capillaries, took approximately 30 minutes for each batch of
samples. The thermal cycler run on the R.A.P.I.D.® 7200 instrument took approximately
30 minutes for each batch of samples. The verification staff analyzed on average three to four
batches of samples on the R. A.P.I.D.® 7200 instrument per day, in some instances analyzing up
to six batches a day. This equated to approximately 36 to 48 duplicate samples being analyzed
per day (72 to 96 actual capillaries loaded) for three to four batches of samples and 72 duplicate
samples a day (144 actual capillaries loaded) for six batches, including two controls (one split
positive and one split negative) for every batch of samples analyzed.
For the purposes of this test, the R. A.P.I.D.® output was monitored using the LCDA
quantification view to determine the results for each sample. The Second Derivative Maximum
function was used to determine the crossing points. Within the LCDA quantification view of the
data, the operator has other options with which he or she can manipulate and inspect the data. A
Fit Points analysis can be used to determine the crossing points for each positive sample. This
option allows for more user-defined criteria for determining the crossing points. For the Fit Points
analysis, the user picks the appropriate baseline adjustment method (as with the Second
Derivative Maximum analysis) and then sets the noise band threshold. The software then
generates the crossing points for each positive sample based on the user-defined criteria. The
R.A.P.I.D.® System also offers a melt cycle and melting curve analysis to aid in product
identification.
Another more simplified screen that was not used in this verification test is also available from
the R. A.P.I.D.® software for interpreting the results. The software also offers a Detector view of
the results at the completion of a PCR run. In this view, all of the samples from a single run on
the R.A.P.I.D.®7200 instrument are shown in one table, including the positive and negative
controls. In this table, the software generates a "Present" response beside a sample when the
bacteria of interest have been detected, and a "Not Detected" beside the sample when the bacteria
have not been detected in the sample. The software also monitors the control samples and will
indicate to the user if a batch of samples needs to be rerun because the controls failed. An
example of the Detector output is shown below in Figure 6-2. An option also exists within the
R. A.P.I.D.® 7200 instrument to test for multiple assays on one carousel batch. Though this option
was not fully verified in this test, for one batch of samples, in an effort to conserve time, two
different assays for the same bacteria (Y. pestis) were analyzed in the same R.A.P.I.D.® 7200
instrument run. No problems were encountered, and the setup was easy.
44
-------
Irucella - Organism Present
*
1
2
3
4
b
6
7
8
9
10
11
12
13
14
15
16
I/
18
19
20
21
22
23
24
2b
l?fi
__ u
27
28
Sample Name
Positive Control
Positive Control
Negative Control
Neqative Control
Sample 7 al
Samp e / a 2
Samp e / bl
Sample / b2
Sample 7 cl
Samp e / c2
Sample 7 dl
Sample 7 c!2
Sample Sal
Samp-C 8 a 2
Samp c 8 bl
Samp e 8 b2
Samp e 8 cl
Sample 8 c2
Sample 8 dl
Sample 8 d2
Sample 9 al
Sample 9 a2
Sample 9 bl
Samp c 9 b2
Samp e 9 cl
Sample 9 c2
Sarnp.e 9 ell
Sample 9 d2
Control
Positive
Positive
Neqatve
Neqat ve
Unknown
Unknown
Unknown
Unknown
Unknown
Ur known
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Unknown
Score
234
195
-8
-8
183
249
127
123
159
132
1/6
212
0
-8
-•:;
-12
— Jj
-8
-5
~ /
-13
-4
-6
-6
-4
.5
_5
-10
Result
Present
Present
Nci Detected
Nci Detected
Present
PreserK
Present
Present
Present
Present
Present
Present
Not Detected
No; Detected
f\c Detected
Nc Detected
l\c Detected
Nc: Detected
Not Detected
No: Detected
Not Detected
Not Detected
Not Detected
Ken Detected
Nci Detected
Nc: Detected
Not Detected
Not Detected
SSN
Enc. 1C
Figure 6-2. R.A.P.LD.® Detector View Output for Spiked (Sample 7) and Unspiked
(Samples 8 and 9) Brucella suis Samples.
45
-------
Chapter 7
Performance Summary
The R.A.P.I.D.® System results for this verification test for samples containing F. tularensis,
Y. pestis, B. anthracis, Brucella suis, and E. coli are presented in Tables 7-1 through 7-9. The
results for each bacteria assay are presented in a separate table. Qualitative responses for each set
of sample replicates as well as accuracy, specificity, false negatives and positives, and precision
are presented in each table. A summary of the other performance factors associated with the
R. A.P.I.D.® System is presented at the end of this chapter. These performance factors apply to the
entire system, across all bacteria.
46
-------
Table 7-1. F. tularensis Target 1 Summary Table
Parameter
Qualitative
results
Contaminant-only
PT samples
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Number Detected/
Sample Information Concentration Number of Samples
4xl05 cfu/mL(a) 4/4
2xl03 cfu/mL 4/4
DI water 5xl03 cfu/mL 4/4
IxlO4 cfu/mL 4/4
5xl04 cfu/mL 4/4
Humic and 1 1A4 , o/i/o/i
,. , . ., 1x10 cfu/mL 24/24
fulvic acids
Concentrated DW 1 x 1 04 cfu/mL 16/16
100% (20 out of 20) of the contaminant-only PT samples above the
system LOD were positive.
88% (21 out of 24) of the unspiked interferent and DW samples were
negative. Three unspiked 2.5 mg/L each humic and fulvic acid
replicates returned inconclusive results.*'
No false positives resulted from the analysis of the unspiked interferent
or DW samples. Three unspiked 2.5 mg/L each humic and fulvic acid
replicates returned inconclusive results.*'
No false negative results were obtained from the analysis of samples
spiked with levels of F. tularensis above the system LOD.
95% (20 out of 21) of the sample sets showed consistent results among
the individual replicates within that set.*'
(a) Infective/lethal dose.
(b) Three unspiked 2.5 mg/L each humic and fulvic acid replicates had one positive and one negative result in the split
samples. These were inconclusive results and would require re-analysis in a real-world scenario. The remaining
replicate was negative.
47
-------
Table 7-2. F. tularensis Target 2 Summary Table
Parameter
Contaminant-
only PT
Qualitative samPles
results
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Sample Information Concentration
4xl05cfu/mL(a)
2xl03 cfu/mL
DI water 5 x 1 03 cfu/mL
IxlO4 cfu/mL
5x1 04 cfu/mL
Humic and 1 1 n4 - , T
,, , . . , lxl04cfu/mL
fulvic acids
Concentrated DW 1 x 1 04 cfu/mL
Number Detected/
Number of Samples
4/4
4/4
4/4
4/4
4/4
23/24(b)
12/16(c)
100% (20 out of 20) of the contaminant-only PT samples above the
system LOD were positive.
96% (23 out of 24) of the unspiked interferent
negative.
and DW samples were
No false positives resulted from the analysis of the unspiked
interferent or DW samples. One unspiked CA DW replicate returned
an inconclusive result. (d)
Four false negative results were obtained from the analysis of samples
spiked with levels of F. tularensis above the system LOD. Three
spiked CA DW and one spiked 0.5 mg/L each humic and fulvic acid
replicates returned negative results.
86% (18 out of 21) of the sample sets showed
among the individual replicates within that set
consistent results
(b, c, d)
(a) Infective/lethal dose.
(b) One spiked 0.5 mg/L each humic and fulvic acid replicate had one negative and three positive results.
(c) One spiked CA DW replicate had one positive and one negative results in the split samples. This indicated an
inconclusive result and would require re-analysis in a real-world scenario. The remaining replicates were negative.
-------
Table 7-3. Y. pestis Target 1 Summary Table
Parameter
Contaminant-
only PT
Qualitative samPles
results
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Sample Information Concentration
0.28 cfu/mL(a)
2xl03 cfu/mL
DI water 5 x 1 03 cfu/mL
IxlO4 cfu/mL
5x1 04 cfu/mL
Humicand 1 1A4 - , T
,, , . . , IxlCrcfu/mL
fulvic acids
Concentrated DW 1 x 1 04 cfu/mL
Number Detected/
Number of Samples
0/4(a)
4/4
4/4
4/4
4/4
24/24
16/16
100% (16 out of 16) of the contaminant-only PT samples above the
system LOD were positive.
100% (24 out of 24) of the unspiked interferent
were negative.
and DW samples
No false positives resulted from the analysis of the unspiked
interferent or DW samples.
No false negative results were obtained from the analysis of the
interferent and DW samples spiked with levels of Y. pestis above the
system LOD.
100% (21 out of 21) of the sample sets showed
among the individual replicates within that set.
consistent results
(a)
Infective/lethal dose—below the R.A.P.I.D.® System LOD for Y. pestis.
49
-------
Table 7-4. Y. pestis Target 2 Summary Table
Parameter
Contaminant-
only PT
Qualitative samPles
results
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Sample Information Concentration
0.28 cfu/mL(a)
2xl03 cfu/mL
DI water 5 x 1 03 cfu/mL
IxlO4 cfu/mL
5x1 04 cfu/mL
Humic and 1 1 n4 , T
,, , . . , IxlCrcfu/mL
fulvic acids
Concentrated DW 1 x 1 04 cfu/mL
Number Detected/
Number of Samples
0/4 (a)
4/4
4/4
4/4
4/4
24/24
16/16
100% (16 out of 16) of the contaminant-only PT samples above the
system LOD were positive.
100% (24 out of 24) of the unspiked interferent and DW samples
were negative.
No false positives resulted from the analysis of the unspiked
interferent or DW samples.
No false negative results were obtained from the analysis of the
interferent and DW samples spiked with levels of Y. pestis above the
system LOD.
100% (21 out of 21) of the sample sets showed
among the individual replicates within that set.
consistent results
(a)
Infective/lethal dose—below the R.A.P.I.D.® System LOD for Y. pestis.
50
-------
Table 7-5. B. anthracis Target 1 Summary Table
Parameter
Qualitative
results
Contaminant-
only PT
samples
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Number Detected/
Sample Information Concentration Number of Samples
200 cfu/mL(a) 0/400
2xl03 cfu/mL 3/4(c)
DI water 5xl03cfu/mL 4/4
IxlO4 cfu/mL 4/4
5x1 04 cfu/mL 4/4
Humic and 1 1A4 , o/i/o/i
,. , . ., 1x10 cfu/mL 24/24
fulvic acids
Concentrated DW lxl04cfu/mL 12/16(4e)
94% (15 out of 16) of the contaminant-only PT samples above the
system LOD were positive.
100% (24 out of 24) of the unspiked interferent and DW samples
were negative.
No false positives resulted from the analysis of the unspiked
interferent or DW samples.
Two false negative results were obtained from the analysis of samples
spiked with levels of B. anthracis above the system LOD, one for
spiked NY DW and the other for spiked CA DW. Inconclusive results
were found for one replicate each for spiked CA DW, spiked FL DW,
and DI water at 2xl03 cfu/mL.(c' ^ e)
76% (16 out of 21) of the sample sets showed consistent results
among the individual replicates within that set.*' c> ^ e)
(a) Infective/lethal dose—below the R.A.P.I.D.® System LOD for B. anthracis.
(b) Three samples in the infective/lethal dose PT sample replicates had one positive and one negative result in the split
samples. This indicated an inconclusive result and would require re-analysis in a real-world scenario. The remaining
replicate was negative.
(c) One PT sample replicate at 2xl03 cfu/mL had one positive and one negative result in the split sample. This indicated
an inconclusive result and would require re-analysis in a real-world scenario. The remaining replicates were positive.
(d) One spiked CA DW replicate had one positive and one negative result in the split sample. This indicated an
inconclusive result and would require re-analysis in a real-world scenario. Two of the remaining replicates were
positive, the other negative.
(e) One spiked FL DW replicate had one positive and one negative result in the split sample. This indicated an
inconclusive result and would require re-analysis in a real-world scenario. Two of the remaining replicates were
positive.
51
-------
Table 7-6. B. anthracis Target 2 Summary Table
Parameter
Qualitative
results
Contaminant-
only PT
samples
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Number Detected/
Sample Information Concentration Number of Samples
200 cfu/mL(a) 0/400
2xl03 cfu/mL 3/4
DI water 5xl03cfu/mL 4/4
IxlO4 cfu/mL 4/4
5x1 04 cfu/mL 4/4
Humic and 1 1A4 , o/i/o/i
,. , . ., 1x10 cfu/mL 24/24
fulvic acids
Concentrated DW lxl04cfu/mL 13/16(c)
94% (15 out of 16) of the contaminant-only PT samples above the
system LOD were positive.
100% (24 out of 24) of the unspiked interferent and DW samples
were negative.
No false positives resulted from the analysis of the unspiked
interferent or DW samples.
Two false negative results were obtained from the analysis of samples
spiked with levels of B. anthracis above the system LOD, one for
spiked CA DW and the other for DI water at 2xl03 cfu/mL.
Inconclusive results were found for two spiked CA DW replicates. (c)
86% (18 out of 21) of the sample sets showed consistent results
among the individual replicates within that set.(b'c)
(a) Infective/lethal dose—below the R.A.P.I.D.® System LOD for B. anthracis.
-------
Table 7-7. B. anthracis Target 3 Summary Table
Parameter
Qualitative
results
Contaminant-
only PT
samples
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Number Detected/
Sample Information Concentration Number of Samples
200 cfu/mL(a) 0/400
2xl03 cfu/mL 4/4
DI water 5xl03cfu/mL 4/4
IxlO4 cfu/mL 4/4
5x1 04 cfu/mL 4/4
Humic and 1 1A4 , o/i/o/i
,. , . ., 1x10 cfu/mL 24/24
fulvic acids
Concentrated DW lxl04cfu/mL 12/16(c)
100% (16 out of 16) of the contaminant-only PT samples above the
system LOD were positive.
100% (24 out of 24) of the unspiked interferent and DW samples
were negative.
No false positives resulted from the analysis of the unspiked
interferent or DW samples.
No false negative results were obtained from the analysis of samples
spiked with levels of B. anthracis above the system LOD. All four
replicates for spiked CA DW returned inconclusive results. (c)
95% (20 out of 21) of the sample sets showed consistent results
among the individual replicates within that set.*'
(a) Infective/lethal dose—below the R.A.P.I.D.® System LOD for B. anthracis.
-------
Table 7-8. Brucella suis Summary Table
Parameter
Contaminant-
only PT
Qualitative samPles
results
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Sample Information
DI water
Humic and
fulvic acids
Concentrated DW
88% (14 out of 16) of the
LOD were positive.
Number Detected/
Concentration Number of Samples
40 cfu/mL(a) 0/4 (b)
2xl03 cfu/mL 2/4 (c)
5xl03 cfu/mL 4/4
IxlO4 cfu/mL 4/4
5 xlO4 cfu/mL 4/4
IxlO4 cfu/mL 24/24
IxlO4 cfu/mL 16/16
contaminant-only PT samples above the system
100% (24 out of 24) of the unspiked interferent and DW samples were
negative.
No false positives resulted from the analysis of the unspiked interferent
or DW samples.
No false negative results were obtained from the analysis of samples
spiked with levels of Brucella suis above the system LOD.
90% (19 out of 21) of the sample sets showed consistent results among
the individual replicates within that set. (b> c)
(a) Infective/lethal dose—below the R.A.P.I.D.® System LOD for Brucella suis.
(b) One sample in the infective/lethal dose PT sample replicates had one positive and one negative result in the split
samples. This indicated an inconclusive result and would require re-analysis in a real-world scenario. The remaining
replicates were negative.
(c) Two PT sample replicate at 2xl03 cfu/mL had one positive and one negative result in the split sample. This indicated
an inconclusive result and would require re-analysis in a real-world scenario. The remaining replicates were positive.
54
-------
Table 7-9. E. coli Summary Table
Parameter
Qualitative
results
Contaminant-
only PT
samples
Interferent
PT samples
DW samples
Accuracy
Specificity
False positives
False negatives
Precision
Sample Information
DI water
Humic and
fulvic acids
Concentrated DW
100% (16 out of 16) of the
system LOD were positive
100% (24 out of 24) of the
negative.
No false positives resulted
or DW samples.
Concentration
0.2 cfu/mL(a)
2xl03 cfu/mL
5xl03 cfu/mL
IxlO4 cfu/mL
5 xlO4 cfu/mL
IxlO4 cfu/mL
IxlO4 cfu/mL
contaminant-only PT
Number Detected/
Number of Samples
0/4
4/4
4/4
4/4
4/4
20/20(b)
16/16
samples above the
unspiked interferent and DW samples were
from the analysis of the unspiked interferent
No false negative results were obtained from the analysis of samples
spiked with levels of E. coli above the system LOD.
100% (20 out of 20) of the sample sets showed consistent results among
the individual replicates within that set.
(a) Infective/lethal dose— below the R.A.P.I.D.® System LOD for E. coli
(b) One set of spiked 0.5 mg/L each humic and fulvic acid replicates were suspected of having cross-contamination
problems. The samples were not rerun, so the results are not reported here for consistency with other rerun reporting.
Other performance factors: A technician with prior PCR experience operated the R.A.P.I.D.®
System at all times. All three components of the R.A.P.I.D.® System (the m 1-2-3 Flow Kit, the
freeze-dried reagents, and the R. A.P.I.D.® 7200 instrument) were straightforward and easy to use.
Three separate work areas were needed for testing to minimize cross-contamination. The freeze-
dried reagents were color coded, contained all of the necessary components for PCR in one vial,
and were reconstituted in the same vial, making PCR setup easy. Reagents for the DNA purifica-
tion and PCR setup had room temperature storage requirements. The glass capillaries used on the
R.A.P.I.D.® 7200 instrument were problematic when the rotors on the carousel were dirty, but
posed no problems once the rotors were properly cleaned. The sample throughput for this
verification test was 36 to 72 samples per day. Approximate operational times were 2 hours for
DNA extraction/purification, 30 minutes for reconstituting the reagents and loading the
capillaries for one carousel batch, and 30 minutes/carousel batch for PCR. The R.A.P.I.D.®
software was easy to use, and additional software analysis tools other than those used in this test
are available. The cost for the R.A.P.I.D.® System is around $8 per sample for the FTI1-2-3 Flow
Kit DNA purification step, $17 to test each split sample using the freeze-dried reagents, and
approximately $55,000 for the R.A.P.I.D.® 7200 instrument itself (20 inches x!4 inches xlO
inches, 50 pounds). The R.A.P.I.D.®7200 instrument can hold up to 32 capillaries (14 split
samples, plus controls).
55
-------
Chapter 8
References
1. Test/QA Plan for Verification of Rapid PCR Technologies, Battelle, Columbus, Ohio,
May 2004.
2. Burrows, W. D.; Renner, S. E. "Biological Warfare Agents as Threats to Potable Water,"
Environmental Health Perspectives, 107,975-984, 1999.
3. Stenman, L; Orpana, A. "Accuracy in amplification," Nature Biotechnology, 19, 1011-
1012,2001.
4. Hughes, L; Totten, P. "Estimating the accuracy of polymerase chain reaction-based tests
using endpoint dilution," University of Washington Biostatistics Working Paper Series,
Working Paper 196, 2003.
5. U.S. EPA Method 180.1, "Turbidity (Nephelometric)" in Methods for the Determination
of Inorganic Substances in Environmental Samples, EPA/600/R-93-100, August 1993.
6. American Public Health Association, et al. Standard Methods for Examination of Water
and Waste\vater. 19th Edition. 1997. Washington D.C.
7. U.S. EPA, Methods for Chemical Analysis of Water and Wastes, EPA/600/4-79-020,
March 1983.
8. U.S. EPA Method 200.8, "Determination of Trace Elements in Waters and Wastes by
Inductively-Coupled Plasma Mass Spectrometry" in Methods for the Determination of
Organic Compounds in Drinking Water, Supplement I, EPA/600/R-94/111, October
1994.
9. U.S. EPA Method 524.2, "Permeable Organic Compounds by Capillary Column
GC/Mass Spectrometry" in Methods for the Determination of Organic Compounds in
Drinking Water, Supplement m, EPA/600/R-95-131. August 1995.
10. U.S. EPA Method 552.2, "Haloacetic Acids and Dalapon by Liquid-Liquid Extraction,
Derivatization and GC with Electron Capture Detector" in Methods for the Determination
of Organic Compounds in Drinking Water, Supplement ffl, EPA/600/R-95-131, August
1995.
56
-------
11. Quality Management Plan (QMP) for the ETV Advanced Monitoring Systems Center,
Version 5.0. EPA Environmental Technology Verification Program, prepared by Battelle,
Columbus, Ohio, March, 2004.
12. Shapiro, D. S. "Quality Control in Nucleic Acid Amplification Methods: Use of
Elementary Probability Theory," Journal of Clinical Microbiology, 37, 848-851, 1999.
57
-------