THE ENVIRONMENTAL TECHNOLOGY VERIFICATION
                                      PROGRAM
     &EPA
                                        Batreiie
                                    The Business of Innovation
                               ETV Verification Statement
   TECHNOLOGY TYPE:   RAPID FUNGI DETECTION

   APPLICATION:          ANALYSIS OF FUNGI IN AIR

   TECHNOLOGY NAME:  Mycometer®-test

   COMPANY:              Mycometer A/S              Mycometer, Inc.
   ADDRESS:
   PHONE:
   WEB SITE:
   E-MAIL:
Lersoe Parkalle 40
2100 Copenhagen
Denmark

+45 3916 1072
www.mycometer.com
info@mycometer.dk
 5002 MacDill Avenue
 Tampa, FL
 33611

 813-434-6998
www.mycometer.com
lrogers@mycometer.com
The U.S. Environmental Protection Agency (EPA) has established the Environmental Technology
Verification (ETV) Program to facilitate the deployment of innovative or improved environmental
technologies through performance verification and dissemination of information. The goal of the ETV
Program is to further environmental protection by accelerating the acceptance and use of improved and cost-
effective technologies. ETV seeks to achieve this goal by providing high-quality, peer-reviewed data on
technology performance to those involved in the design, distribution, financing, permitting, purchase, and
use of environmental technologies. Information and ETV documents  are available at www.epa.gov/etv.

ETV works in partnership with recognized standards and testing organizations, with stakeholder groups
(consisting of buyers, vendor organizations, and permitters), and with individual technology developers. The
program evaluates the performance of innovative technologies by developing test plans that are responsive to
the needs of stakeholders, conducting field and laboratory tests (as appropriate), collecting and analyzing
data, and preparing peer-reviewed reports. All evaluations are conducted in accordance with rigorous quality
assurance (QA) protocols to ensure that data of known and adequate quality are generated and that the results
are defensible.

The Advanced Monitoring Systems (AMS) Center, one of six verification centers under ETV, is operated by
Battelle in cooperation with EPA's National Risk Management Research Laboratory. The AMS Center
evaluated the performance of a rapid fungi detection technology. This verification statement provides a
summary of the test results for Mycometer®-test developed by Mycometer A/S and distributed in the United
States by Mycometer, Inc.

-------
VERIFICATION TEST DESCRIPTION

Rapid technologies (results available same day of testing) to detect fungi from matrices such as surfaces,
bulk material, air, or water are of interest to improve the efficiency of delineating and documenting fungal
contamination in buildings and water systems, and for monitoring progress during cleanup and remediation
processes. Traditional methods of analysis can take up to seven days for results. Technologies providing
same day or near "real-time" results indicating changes  in air quality would help to control diseases
associated with fungal outbreaks, expedite remediation efforts, and protect public health.

The verification test of the Mycometer®-test technology was conducted from May 19 through June 9, 2011 at
Battelle in Columbus, Ohio.  Technology operation, sample handling, and analyses were performed
according to the vendor's instructions.

For this verification, the Mycometer®-test technology was verified for repeatability and inter-assay
reproducibility by detecting fungi in air samples. Linearity was assessed using dilutions of stock cultures in
dechlorinated tap water. The linearity test for fungi was a modification of test procedures in place for air and
surface samples. In addition, sustainable operational factors such as ease of use, required reagents, analysis
time, laboratory space, and utilities required were reported.

QA oversight of verification testing was provided by Battelle and EPA. Battelle and EPA QA staff
conducted technical systems audits of the testing and Battelle QA staff conducted a data quality audit of at
least 10% of the test data. This verification statement, the full report on which it is based, and the quality
assurance project plan for this verification test are available at www.epa.gov/etv/centers/center 1 .html.

TECHNOLOGY DESCRIPTION

The Mycometer®-test rapid fungi detection technology is based on fluorogenic detection of enzyme activities
found predominantly in a taxonomic  group of organisms. A sample (e.g., filter or swab) is contacted with a
test solution containing a synthetic enzyme substrate. The enzyme present in the fungal cells hydrolyzes the
synthetic enzyme substrate.  When the  synthetic substrate molecule is cleaved into two molecules by the
enzyme, one of the molecules can be made to fluoresce upon excitation with ultraviolet (UV) light (365
nanometers). The amount of fluorescence is measured using a handheld fluorometer after processing for a
reaction time based on the ambient temperature.  This fluorescence semi-quantitatively correlates to a
measure of the fungal biomass. Fluorescence measurements can be captured electronically and may be
downloaded to a computer or can be transcribed by hand. The sample preparation and analyses can be
performed on site in less than one hour.

According to the vendor, the Mycometer®-test for fungi is designed to measure both viable and non-viable
spores, hyphae and fungal particles such as hyphal fragments in air, on surfaces, or in bulk materials to give a
representation of the contamination in the environment. Although the Mycometer®-test cannot distinguish
between fungal genera or viable/non-viable fungi, it provides a semi-quantitative measure of the total fungal
biomass present. Air samples can be collected with traditional air sampling pumps onto filter media.
Typically 300 liters of air are collected by sampling 20 liters per minute (LPM) for 15 minutes, or 15 LPM for
20 minutes.  Surface samples are collected by swabbing a nine square centimeter area and bulk material
samples are weighed. Enzyme substrate is added to the filter, swab, or bulk material and the fungal enzyme
reacts with the substrate to release  a fluorescent product. The amount of fungi in the sample is estimated by
measuring the fluorescence produced.  For the Mycometer®-test, the type and material of the air sampling
filter cartridge are critical for both  sampling and the enzyme reaction (which take place directly on the filter).
It is important, therefore, to  use the filters provided by the vendor.  The vendor provides a proficiency
certification training program that is included with the fluorometer kit (on a flash drive) and is mandatory for
use of their technology to document understanding and proper training.

-------
VERIFICATION RESULTS
Table 1 summarizes the linearity results for Mycometer®-test using two fungal cultures in water, Aspergillus
flavus ATCC 58870 and Cladosporium herbarum ATCC 58927.

Table 1. Linearity Results for Mycometer®-test Adjusted Fluorescence vs. Total Spores Tested
Test
Organism
A. flavus
ATCC 58870
C. herbarum
ATCC 58927
Total Spores
Tested
3.1xl05to6.2xl06
4.8xl05to9.6xl06
Range of
Average
Adjusted
Fluorescence
Units (fu)
218 to 4322
125 to 3389
Slope
0.0007
0.0004
Y-
intercept
20.637
-135.25
Coefficient of
Determination
(R2)
0.9979
0.9976
Adjusted fluorescence = sample fluorescence reading - blank fluorescence reading

Table 2 summarizes the repeatability results for Mycometer®-test using eight replicates of one fungal culture
in air, all analyzed by one person.

Table 2. Mycometer®-test Repeatability: Air Containing A Flavus
Test Iteration
Average
Standard Deviation
RSD (%)
Adjusted Fluorescence Units (fu)
A. flavus (6. 2 xlO3
n=8
spore s/L)
334
27
8.0
Table 3 summarizes the inter-assay reproducibility results for Mycometer®-test using eight replicates of one
fungal culture in air split into four samples each for analysis by two people with two different fluorometers.

Table 3. Mycometer®-test Inter-Assay Reproducibility: Air Containing A Flavus
Test Iteration
Average
Standard Deviation
RSD (%)
RPD (%)
Adjusted Fluorescence Units (fu)
A. flavus (6.2 x 103 spores/L)
n=4
Analyst 1
291
25
8.7
Analyst 2
307
15
4.7
5.3
Operational Factors. The verification staff found that the Mycometer®-test was easy to use.  A Mycometer
A/S representative came to Battelle to train the verification staff in the use of the Mycometer®-test reagents
and operation of the fluorometer. This training lasted one day and staff felt it was more than sufficient to be
comfortable using the reagent kits and fluorometer without assistance.  This on-site training focused on the
technology operating protocols for air and water matrices. While the operational aspects of this training
were similar to the proficiency certification program, the proficiency certification program also focuses on
understanding the principles behind the technology as well as additional applications.

-------
The fluorometer is provided in a hard-cover carrying case.  The carrying case has dimensions of 45 cm wide
x 15 cm deep * 32 cm high (17.5 in wide x 6 in deep  * 12.5 in high) and weighs approximately 7.2
kilograms (16 pounds). Included with the fluorometer is a black calibration cuvette, a 100 uL automatic
pipette, a timer, two test racks, a calculator, a thermometer, and training materials. The fluorometer operates
on four AAA batteries and has push-button operation. Testing staff found that the display was easy to read
and surfaces were easy to wipe clean. The fluorometer required a calibration check once daily with the black
cuvette provided with the fluorometer and a calibration standard provided in the reagent kit.  Both an
instruction manual and a quick reference card were provided for the Mycometer®-test. Verification staff
found that the instructions provided were not always consistent between the manual and the quick reference.
For example, the manual indicated that the blank sample for air testing was to be a blank filter processed
alongside the test filters, while the quick reference guide indicated that the blank was to be an aliquot of the
substrate combined with the developer.

The Mycometer®-test reagents are sold in lots of 10 for air assays and lots of 20 for surface assays. Each
reagent kit included the sampling media (filter for air  samples), enzyme substrate, developer, and calibration
standard, all of which were clearly labeled for identification and storage conditions. Syringes and cuvettes
used for processing were also included.  All containers and packaging were easy to open; however,
verification staff found there was packaging waste involved with the different components, particularly if
multiple kits were needed to analyze the required number of samples. All reagents were ready for use with
the exception of the enzyme substrate which required  re-hydration.  Each sample resulted in approximately 5
mL of liquid waste from the substrate and developer used to process the sample. Based on the expiration
date stamped on the kits, the shelf life of the  kits received for testing was over one year from receipt date.
Several kit components required refrigeration.  Once rehydrated, the enzyme substrate could be stored in a
refrigerator for up to one week or at -18 °C for up to 6 months. All components needed to prepare  and
analyze a sample were included either in the  reagent kit or the fluorometer kit.  No other laboratory
equipment was needed for processing air samples. For air sample collection, however, a sampling pump
must be obtained.  The recommended air sampling pumps (Gast 3-30 LPM IAQ Pump w/Tubing &
Rotameter) are commercially available.  Pricing for the fluorometer and reagent kits can be obtained from
the provided vendor contacts.  Verification testing staff found they were  able to collect and analyze eight air
samples in one hour, given the availability of enough  air sampling pumps to generate eight air samples
simultaneously.

For data reduction, a laptop or personal computer is needed. Mycometer  provides an Excel spreadsheet for
quantification of mold/fugi in  air that converts fluorescence unit values into a 'Mycometer-Air' value and
provides suggested interpretation guidelines based on the resulting value obtained. The Mycometer- Air
value calculation converts the  fluorescence reading to fluorescence units (fu) per volume of air measured in
cubic meters. This can be used to standardize the results for consistent comparison and interpretation if there
are slight variations in the air volume sampled. Because all sample volumes used in verification testing were
the same, conversion of results to a Mycometer-Air value were not needed for verification testing. In
addition, the interpretation guidelines associated with  the Mycometer-Air values were not verified as part of
this test.
Signed by Tracy Stenner	01/06/2012   Signed by Cynthia Sonich-Mullin  02/01/2012
Tracy Stenner                         Date         Cynthia Sonich-Mullin                    Date
Manager                                            Director
Environmental Solutions Product Line                 National Risk Management Research Laboratory
Energy, Environment, and Material Sciences            Office of Research and Development
Battelle                                             U.S. Environmental Protection Agency

-------