THE ENVIRONMENTAL TECHNOLOGY VERIFICATION
PROGRAM
oEPA
U.S. Environmental Protection Agency
* J NSF International
ETV Joint Verification Statement
TECHNOLOGY TYPE: POINT-OF-USE REVERSE OSMOSIS DRINKING WATER
TREATMENT SYSTEM
APPLICATION: REMOVAL OF MICROBIAL CONTAMINATION AGENTS IN
DRINKING WATER
PRODUCT NAME: WATTS PREMIER ULTRA 5
COMPANY: WATTS PREMIER, INC.
ADDRESS: 1725 WEST WILLIAMS STREET PHONE: 800-752-5582
PHOENIX, AZ 85027 FAX: 623-931-0191
NSF International (NSF) manages the Drinking Water Systems (DWS) Center under the U.S.
Environmental Protection Agency's (EPA) Environmental Technology Verification (ETV) Program. The
DWS Center recently evaluated the performance of the Watts Premier, Inc. Ultra 5 point-of-use (POU)
reverse osmosis drinking water treatment system. NSF performed all of the testing activities, and also
authored the verification report and this verification statement. The verification report contains a
comprehensive description of the test.
EPA created the ETV Program to facilitate the deployment of innovative or improved environmental
technologies through performance verification and dissemination of information. The goal of the ETV
program is to further environmental protection by accelerating the acceptance and use of improved and
more cost-effective technologies. ETV seeks to achieve this goal by providing high-quality, peer-
reviewed data on technology performance to those involved in the design, distribution, permitting,
purchase, and use of environmental technologies.
ETV works in partnership with recognized standards and testing organizations, stakeholder groups
(consisting of buyers, vendor organizations, and permitters), and with the full participation of individual
technology developers. The program evaluates the performance of innovative technologies by developing
test plans that are responsive to the needs of stakeholders, conducting field or laboratory tests (as
appropriate), collecting and analyzing data, and preparing peer reviewed reports. All evaluations are
conducted in accordance with rigorous quality assurance protocols to ensure that data of known and
adequate quality are generated and that the results are defensible.
NSF 04/12/EPADWCTR The accompanying notice is an integral part of this verification statement. June 2004
VS-i
-------
ABSTRACT
The Watts Premier Ultra 5 was tested for removal of bacteria and viruses at NSF's Drinking Water
Treatment Systems Laboratory. Watts Premier submitted ten units, which were split into two groups of
five. One group received 25 days of conditioning prior to challenge testing, while the second group was
tested immediately. Due to an incorrectly installed shut-off valve on one of the unconditioned units, only
four in this group were tested. Both groups were challenged identically. The challenge organisms were
the viruses fr, MS2, and Phi X 174, and the bacteria Brevundimonas diminuta and Hydrogenophaga
pseudoflava. The test units were challenged at two different inlet pressures - 40 and 80 pounds per
square inch, gauge (psig). The virus challenges were conducted at three different pH settings (6, 7.5, and
9) with the intent to assess whether pH influenced the performance of the test units. The bacteria
challenges were only conducted at pH 7.5.
In most cases, the test units significantly reduced the challenge organisms, with reductions greater than
4.0 logic. The logio reduction data is shown in Tables 3 through 6. Overall, the performance of the
conditioned units was better than that of the unconditioned units. Also, the unconditioned units exhibited
wider unit-to-unit performance variation than the conditioned units. The logio reduction data does not
conclusively show that inlet pressure or pH influenced test unit performance.
TECHNOLOGY DESCRIPTION
The following technology description was provided by the manufacturer and has not been verified.
The Watts Premier Ultra 5 is a five-stage POU drinking water treatment system. It employs carbon
filtration and reverse osmosis processes to remove contaminants from drinking water. It is sold with a
faucet that is installed at the kitchen sink, and the system itself is installed either under the kitchen sink or
in another location.
During operation, inlet water first passes through a sediment filter, and then through two carbon block
filters. The fourth stage is passage through the reverse osmosis membrane. The portion of the inlet water
that passes through the membrane travels to the product water storage tank. When the user opens the
faucet, the water leaves the storage tank and travels through a final carbon filter before exiting the faucet.
The system is designed to produce approximately 12 gallons of reject water for each gallon of treated
water produced.
The test units were evaluated without the carbon filters or sediment filter in place to eliminate the
possibility that these filters could temporarily trap a portion of the challenge organisms, causing a positive
bias of system performance during testing.
VERIFICATION TESTING DESCRIPTION
Test Site
The testing site was the Drinking Water Treatment Systems Laboratory at NSF in Ann Arbor, Michigan.
A description of the test apparatus can be found in the test/quality assurance (QA) plan and verification
report. The testing was conducted in September and October of 2003.
Methods and Procedures
The testing methods and procedures are detailed in the Test/QA Plan for Verification Testing of the Watts
Premier Ultra 5 Point-of-Use Reverse Osmosis Drinking Water Treatment System for Removal of
Microbial Contamination Agents. Nine test units were verified for bacteria and virus removal
NSF 04/12/EPADWCTR The accompanying notice is an integral part of this verification statement. June 2004
VS-ii
-------
performance using the bacteriophage viruses fr, MS2, and Phi X 174, and the bacteria B. diminuta and H.
pseudoflava. The challenge organisms were chosen because they are smaller than most other viruses and
bacteria, and so provide a conservative estimate of performance.
Watts Premier submitted ten units, which were split into two groups of five according to the performance
of each membrane in the manufacturer's quality control testing. One group was conditioned for 25 days
prior to challenge testing by operating the units daily using the test water without challenge organisms.
The second group was challenged without receiving the 25-day conditioning period. Due to an
incorrectly installed shut-off valve on one of the unconditioned units, only four in this group were tested.
The test units were challenged at both 40 and 80 psig inlet pressure. The test water for the bacteria
challenges was set to pH 7.5 ± 0.5. The test water for the virus challenges was set at pH 6.0 ± 0.5, 7.5 ±
0.5, and 9.0 ± 0.5. However, it had a low buffering capacity, so the lab technicians had difficulty
maintaining the pH within the 9.0 ± 0.5 range. As a result, the pH for the conditioned units pH 9, 80 psig
challenge was only 7.9. The test water pH values for all other challenges were within the allowable
ranges. These challenge conditions were intended to evaluate whether inlet pressure or pH influences
bacteria and virus removal. Table 1 shows the challenge schedule for the conditioned units, while Table 2
shows the schedule for the unconditioned units. The challenge levels ranged from 3.4 to 6.4 logic for the
viruses, and 6.7 to 8.4 logio for the bacteria.
Table 1. Conditioned Units Challenge Schedule
pH Inlet Pressure
Day Challenge Organism(s) (±0.5 units) (± 3 psig)
1
2
3
4
5
6
7
8
9
10
Day
1
2
3
4
5
6
7
8
9
10
All Viruses
All Viruses
All Viruses
All Viruses
All Viruses
All Viruses
H. pseudoflava
H. pseudoflava
B. diminuta
B. diminuta
Table 2. Unconditioned
Challenge Organism(s)
H. pseudoflava
H. pseudoflava
B. diminuta
B. diminuta
All Viruses
All Viruses
All Viruses
All Viruses
All Viruses
All Viruses
6.0
6.0
7.5
7.5
9.0
9.0
7.5
7.5
7.5
7.5
Units Challenge
pH
(±0.5 units)
7.5
7.5
7.5
7.5
6.0
6.0
7.5
7.5
9.0
9.0
40
80
40
80
40
80
80
40
40
80
Schedule
Inlet Pressure
(± 3 psis)
80
40
40
80
40
80
40
80
40
80
NSF 04/12/EPADWCTR The accompanying notice is an integral part of this verification statement. June 2004
VS-iii
-------
On each challenge day, the test units were operated for one tank-fill period (approximately six to eight
hours). The end of this period was evident through engagement of the system's automatic shutoff
mechanism, which causes the flow of reject water to cease. At 40 psig, not all of the shut-off mechanisms
engaged after 8 hours of operation due to the low pressure. The storage tanks were nearly full in these
instances, so operation of the units was stopped manually.
Influent water samples were collected at the beginning and end of the challenge period. After each test
unit ceased operation, the entire contents of the product water storage tank were emptied into a sterile
container, and a subsample was collected for microbiological analysis. All samples were enumerated in
triplicate. Following each challenge period, the test units were flushed by operating them for one tank-fill
period using the test water without challenge organisms.
VERIFICATION OF PERFORMANCE
The bacteria reduction data are presented in Tables 3 and 4, and the virus reduction data in Tables 5 and
6. An examination of the bacteria reduction data shows that for the five conditioned test units, in only
one case (unit 4 for B. diminuta at pH 7.5, 40 psig) was one of the bacteria species detected in the effluent
samples. In contrast, for the unconditioned units, there were 13 cases out of 16 where the challenge
bacteria were detected in the effluents.
An evaluation of the virus reduction data shows that overall, the conditioned units performed better than
the unconditioned units. The mean logio reductions and mean logio effluent counts are shown in the
bottom right corner of Tables 5 and 6. A comparison of the mean logio effluent counts for the
unconditioned versus conditioned units shows that the conditioned units performed approximately 0.3 to
1.7 logio better than the unconditioned units.
The unit-to-unit performance variation for the unconditioned units was wider than for the conditioned
units, and the performance of each unconditioned unit also varied more from day-to-day. Also, the
unconditioned units had many cases where bacteria reduction performance was less than virus reduction
performance. The reasons for these observations are not known, but the data suggest that conditioning the
systems improves and/or stabilizes their performance. The data does not conclusively show whether inlet
pressure or pH influenced test unit performance.
Table 3. Bacteria Log Reduction Data for Unconditioned Units
Pressure Challenge Log10 Influent Geometric Mean Log10 Reduction
pH (psig) Organisms Challenge Unit 1 Unit 2 Unit 3 Unit 4
7.5
40
H. pseudoflava
B. diminuta
6.9
8.2
4.4
8.2
4.9
3.0
2.2
2.0
1.6
8.2
7.5
80
H. pseudoflava
B. diminuta
6.9
8.1
4.6
3.5
6.6
2.2
1.9
3.3
3.0
8.1
Table 4. Bacteria Log Reduction Data for Conditioned Units
Pressure Challenge Log10 Influent Geometric Mean Log10 Reduction
pH (psig) Organisms Challenge Unit 1 Unit 2 Unit 3 Unit 4 Unit 5
~1~5 40 H. pseudoflava 6J 6J 6J 6J 6J 6.7
B. diminuta 8.3 8.3 8.3 8.3 7.2 8.3
7.5
80
H. pseudoflava
B. diminuta
6.7
8.4
6.7
8.4
6.7
8.4
6.7
8.4
6.7
8.4
6.7
8.4
NSF 04/12/EPADWCTR The accompanying notice is an integral part of this verification statement. June 2004
VS-iv
-------
Table 5. Virus Log Reduction
Challenge Conditions Logic
Target Actual Pressure Challenge Influent
Data for Unconditioned Units
Geometric Mean Logic Reduction
pH pH (psig) Organisms Challenge Unit 1 Unit 2 Unit 3
6.0 ±0.5 6.5 40 fr
MS2
Phi X 174
6.0 ±0.5 6.2 80 fr
MS2
Phi X 174
7.5 ±0.5 7.6 40 fr
MS2
Phi X 174
7.5 ±0.5 7.7 80 fr
MS2
Phi X 174
9.0 ±0.5 8.7 40 fr
MS2
Phi X 174
9.0 ±0.5 9.0 80 fr
MS2
Phi X 174
6.3
6.1
5.0
5.9
5.8
4.9
5.9
5.6
5.7
5.8
5.7
5.9
5.8
5.6
5.7
6.0
5.7
5.6
fr mean3
MS2 mean3
PhiX
174 mean3
4.8
5.62
5.0
4.5
4.5
4.62
4.0
3.8
3.7
4.6
4.4
4.3
4.4
4.1
3.8
4.6
4.7
4.1
4.5
4.5
4.3
3.1
3.0
2.4
3.2
3.0
2.8
2.9
2.7
2.3
2.5
2.6
2.6
2.9
2.7
2.6
3.5
3.4
3.5
3.0
2.9
2.7
2.9
2.8
2.3
3.3
3.3
2.4
4.9
5.0
5.72
4.3
4.3
3.7
4.2
4.1
3.3
3.7
3.8
3.5
3.9
3.9
3.5
Unit 4
4.6
4.7
5.02
5.9
5.8
4.9
4.4
4.3
4.3
5.5
5.42
5.1
4.8
4.8
4.1
5.1
5.1
4.5
5.1
5.0
4.7
Mean1
3.8
4.0
3.7
4.2
4.2
3.7
4.1
4.0
4.0
4.2
4.2
3.9
4.1
3.9
3.5
4.2
4.3
3.9
4.1
4.1
3.6
Log10
Mean
Effluent
Count
2.5
2.1
1.3
1.7
1.6
1.2
1.8
1.6
1.7
1.6
1.5
2.0
1.7
1.7
2.2
1.8
1.4
1.7
1.9
1.7
1.7
1 The arithmetic mean of all test units for each challenge.
2 Triplicate count had two
"non-detect"
3 The arithmetic mean for all challenges
agar plates.
against
each test unit.
NSF 04/12/EPADWCTR
The accompanying notice is an integral part of this verification statement.
VS-v
June 2004
-------
Table 6. Virus Log Reduction Data for Conditioned Units
Challenge Conditions Log10 Geometric Mean Logic Reduction
Target Actual Pressure Challenge Influent
pH pH (psig) Organisms Challenge Unit 1 Unit 2 Unit 3 Unit 4 Unit 5
6.0 ±0.5 6.5 40 fr 5.1 3.6 4.1 4.0
MS2 4.8 3.2 3.7 3.8
Phi X 174 3.4 3.4 3.4 3.4
6.0 ±0.5 6.4 80 fr 6.1 4.6 4.2 4.3
MS2 6.0 4.6 4.2 4.2
Phi X 174 3.8 3.8 3.8 3.8
7.5 ±0.5 7.5 40 fr 6.4 4.2 4.8 4.7
MS2 6.2 4.2 4.5 4.8
Phi X 174 4.0 3.7 4.02 4.02
7.5 ±0.5 7.3 80 fr 6.3 4.8 5.6 5.6
MS2 6.1 5.2 5.5 5.6
PhiX174 4.1 4.1 4.12 4.1
9.0 ±0.5 8.9 40 fr 6.2 4.4 4.2 4.3
MS2 5.8 4.2 4.0 4.2
PhiX174 4.1 4.1 4.1 4.1
9.0 ±0.5 7.93 80 fr 6.0 4.4 4.9 4.7
MS2 5.9 4.3 5.9 4.8
Phi X 174 4.0 4.0 4.0 4.0
frmean4 4.3 4.6 4.6
MS2mean4 4.3 4.6 4.6
Phi X 174 mean4 3.9 3.9 3.9
1 The arithmetic mean of all test units for each challenge.
2 Triplicate count had two "non-detect" agar plates.
4.8
4.1
3.4
4.7
4.8
3.8
4.8
4.7
4.0
5.3
4.9
4.1
4.3
4.1
4.1
4.7
4.9
4.0
4.8
4.6
3.9
4.0
3.2
3.4
4.6
3.7
3.8
4.2
4.3
3.7
4.8
5.0
4.12
4.3
4.2
4.1
4.6
4.6
4.0
4.4
4.2
3.9
Mean1
4.1
3.6
3.4
4.5
4.3
3.8
4.5
4.5
3.9
5.2
5.2
4.1
4.3
4.1
4.1
4.7
4.9
4.0
4.6
4.4
3.9
Mean
Effluent
Count
1.0
1.2
0.0
1.6
1.7
0.0
1.9
1.7
0.1
1.1
0.9
0.1
1.9
1.7
0.0
1.3
1.0
0.0
1.5
1.4
0.0
3 See section 5.8.3 of verification report for discussion of pH variance.
4 The arithmetic mean for all challenges against each test unit.
QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)
NSF personnel conducted a technical systems audit during testing to
compliance with the test plan. NSF also conducted a data quality audit
the
verification report referenced below for more QA/QC information.
ensure
that
the testing was in
of 100% of the data.
Please see
NSF 04/12/EPADWCTR
The accompanying notice is an integral part of this verification statement.
VS-vi
June 2004
-------
Original signed by Original signed by
E. Timothy Oppelt 07/12/04 Gordon Bellen 07/16/04
E. Timothy Oppelt Date Gordon Bellen Date
Director Vice President
National Homeland Security Research Center Research
United States Environmental Protection Agency NSF International
NOTICE: Verifications are based on an evaluation of technology performance under specific,
predetermined criteria and the appropriate quality assurance procedures. EPA and NSF make no expressed
or implied warranties as to the performance of the technology and do not certify that a technology will
always operate as verified. The end user is solely responsible for complying with any and all applicable
federal, state, and local requirements. Mention of corporate names, trade names, or commercial products
does not constitute endorsement or recommendation for use of specific products. This report is not a NSF
Certification of the specific product mentioned herein.
Availability of Supporting Documents
Copies of the test protocol, the Verification Statement, and the Verification Report (NSF Report # NSF
04/12/EPADWCTR) are available from the following sources (NOTE: Appendices are not included in the
Verification Report. Appendices are available from NSF upon request.):
1. ETV Drinking Water Systems Center Manager (order hard copy)
NSF International
P.O. Box 130140
Ann Arbor, Michigan 48113-0140
2. NSF web site: http://www.nsf.org/etv/dws/dws reports.html and from
http://www.nsf.org/etv/dws/dwsjroject documents.html (electronic copy)
3. EPA web site: http://www.epa.gov/etv (electronic copy)
NSF 04/12/EPADWCTR The accompanying notice is an integral part of this verification statement. June 2004
VS-vii
------- |