EPA/600/R-08/101
METHOD 522 DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY
            SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/
            MASS SPECTROMETRY (GC/MS) WITH SELECTED ION
            MONITORING (SIM)
                               Version 1.0
                             September, 2008
                    Jean W. Munch and Paul E. Grimmett
             NATIONAL EXPOSURE RESEARCH LABORATORY
                OFFICE OF RESEARCH AND DEVELOPMENT
              U. S. ENVIRONMENTAL PROTECTION AGENCY
                        CINCINNATI, OHIO 45268
                                 522-1

-------
                                    METHOD 522

   DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE
   EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY
                (GC/MS) WITH SELECTED ION MONITORING (SIM)
1.  SCOPE AND APPLICATION

   1.1   This is a gas chromatography/mass spectrometry (GC/MS) method for the
         determination of 1,4-dioxane (CASRN 123-91-1) in drinking water.  Accuracy and
         precision data have been generated in reagent water, finished ground and surface
         waters. Although this method was developed and demonstrated using selected ion
         monitoring (SIM) GC/MS for maximum sensitivity, it can also be used with full scan
         GC/MS if the sensitivity achieved meets the user's data requirements.

   1.2   The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets
         Data Quality Objectives (DQOs) that are developed based on the  intended use of this
         method. Two single laboratory Lowest Concentration Minimum  Reporting Levels
         (LCMRLs) of 0.036 and 0.047 ug/L have been determined in reagent water.  The
         single laboratory LCMRL is the lowest true concentration for which the future
         recovery is predicted to fall, with high confidence (99%), between 50 and 150%
         recovery. The procedure used to determine the LCMRLs is described elsewhere.1

   1.3   Laboratories using this method will not be required to determine the LCMRL for this
         method, but will need to demonstrate that their laboratory MRL for this method meets
         the requirements described in Section 9.2.4.

   1.4   Determining the Detection Limit (DL) for analytes in this method is optional (Sect.
         9.2.6).  Detection limit is defined as the statistically calculated minimum concentration
         that can be measured with 99% confidence that the reported value is greater than zero.2
         The DL is compound dependent and is dependent on extraction efficiency, sample
         matrix, fortification concentration, and instrument performance.

   1.5   This method is intended for use only by analysts skilled in solid phase extraction
         (SPE), the operation of GC/MS instruments, and the interpretation of the associated
         data.

   1.6   METHOD FLEXIBILITY - In recognition of technological advances in analytical
         systems and techniques, the laboratory is permitted to modify the GC column, GC
         conditions and MS conditions (Sect.  9.4). Changes may not be made to sample
         collection and preservation (Sect. 8), the quality control (QC) requirements (Sect. 9),
         or to the sample extraction steps (Sect. 11). Method modifications should be
         considered only to improve method performance. Modifications that are introduced in
         the interest of reducing cost or sample processing time, but result in poorer method
         performance, should not be used. For example, modifications should not sacrifice

                                        522-2

-------
         chromatographic separations in the interest of method turnaround time. Because the
         method analyte, surrogate analyte (SUR) and internal standard (IS) have only very low
         mass ions in their mass spectra, and SIM is being used, it is important to strive for
         chromatographic separation from potential interferences. In all cases where method
         modifications are proposed, the analyst must perform the procedures outlined in the
         initial demonstration of capability (IDC; Sect. 9.2), verify that all QC acceptance
         criteria in this method (Sect. 9) are met, and demonstrate that method performance can
         be verified in a real sample matrix (Sect. 9.3.7 and 9.4).

         NOTE: The above method flexibility section is intended as an abbreviated summation
         of method flexibility.  Method Sections 4-12 provide detailed information of specific
         portions of the method that may be modified. If there is any perceived conflict
         between the general method flexibility statement in this section and specific
         information in Sections 4-12, Sections 4-12 supersede Section 1.6.

2. SUMMARY OF METHOD

   2.1   A water sample that has been dechlorinated and preserved with a microbial inhibitor is
         fortified with the isotopically labeled SUR, 1,4-dioxane-t/g. The sample is extracted
         by one of two SPE options.  In option 1, a 500-mL sample is passed through an SPE
         cartridge containing 2 g of coconut charcoal to extract the method analyte and SUR.
         In option 2, a 100-mL sample is extracted on a Waters AC-2 Sep-Pak or Supelco
         Supelclean ENVI-Carb Plus cartridge.  In either option, the compounds are eluted
         from the solid phase with a small amount of dichloromethane (DCM), approximately
         9 mL or 1.5 mL, respectively. The extract volume is adjusted, and the IS,
         tetrahydrofuran-J8 (THF-J8), is added. Finally, the extract is dried with anhydrous
         sodium sulfate. Analysis of the extract is performed by GC/MS. The data provided in
         this method were collected using splitless injection with a high-resolution fused silica
         capillary GC column that was interfaced to an MS operated in the SIM mode.  The
         analyte, SUR and IS are separated  and identified by comparing the acquired mass
         spectra and retention times to reference spectra and retention times for calibration
         standards  acquired under identical  GC/MS conditions.  The concentration of the
         analyte is  determined by comparison to its response in calibration standards relative to
         the IS.  Although the performance  data presented in Section 17 of this method were
         obtained in the SIM mode for maximum method sensitivity, the sample extract
         analysis may also be performed in  full scan mode if the sensitivity achieved meets the
         data user's requirements.

3. DEFINITIONS

   3.1   ANALYSIS BATCH - A set of samples that is analyzed on the same instrument
         during a 24-hour period that begins and ends with the analysis of the appropriate
         Continuing Calibration Check (CCC) Standards.  Additional CCCs may be required
         depending on the length of the analysis batch and/or the number of Field Samples.
                                         522-3

-------
3.2   CALIBRATION STANDARD (CAL) - A solution prepared from stock standard
      solution(s) and the ISs and SURs. The CAL solutions are used to calibrate the
      instrument response with respect to analyte concentration.

3.3   CONTINUING CALIBRATION CHECK (CCC) STANDARD - A calibration
      standard containing the method analyte, IS and SUR which is analyzed periodically to
      verify the accuracy of the existing calibration for those analytes.

3.4   DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be
      identified, measured and reported with 99% confidence that the analyte concentration
      is greater than zero.  This is a statistical determination (Sect. 9.2.6), and accurate
      quantitation is not expected at this level.2

3.5   EXTRACTION BATCH - A set of up to 20 Field Samples (not including QC
      samples) extracted together by the same person(s) during a work day using the same
      lot of SPE devices, solvents, SUR solution, and fortifying solutions.  Required QC
      samples include Laboratory Reagent Blank (LRB), Laboratory Fortified Blank (LFB),
      Laboratory Fortified Sample Matrix (LFSM), and either a Field Duplicate (FD1 and
      FD2) or Laboratory Fortified Sample Matrix Duplicate (LFSMD).

3.6   FIELD  DUPLICATES (FD1 and FD2) - Two separate samples collected at the same
      time and place under identical circumstances, and treated exactly the same throughout
      field and laboratory procedures.  Analyses of FD1 and FD2 give a measure of the
      precision associated with sample collection, preservation, and storage, as well as with
      laboratory procedures.

3.7   INTERNAL STANDARD (IS) - A pure analyte, which is extremely unlikely to be
      found in any sample, which is added to an extract or standard solution in a known
      amount and used to measure the relative responses of the method analyte and SUR.
      The IS must be an analyte that is not a sample component.

3.8   LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other
      blank matrix to which known quantities of the method analyte and all the preservation
      compounds are added. The LFB is processed and analyzed exactly like a sample, and
      its purpose is to determine whether the methodology is in control, and whether the
      laboratory is capable of making accurate and precise measurements.

3.9   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a Field
      Sample to which known quantities of the method analyte and all the  preservation
      compounds are added. The LFSM is  processed and analyzed exactly like a sample,
      and its purpose is to determine whether the sample matrix contributes bias to the
      analytical results.  The background concentrations of the analyte in the sample matrix
      must be determined in a separate aliquot or duplicate sample and the measured values
      in the LFSM corrected for background concentrations.
                                     522-4

-------
3.10  LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A Field
      Sample Duplicate of the Field Sample used to prepare the LFSM, which is fortified,
      extracted and analyzed identically to the LFSM. The LFSMD is used instead of the
      Field Duplicate to access method precision and accuracy when the occurrence of the
      method analyte at a concentration greater than the MRL is infrequent.

3.11  LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other
      blank matrix that is treated exactly as a sample including exposure to all glassware,
      equipment, solvents, reagents, sample preservatives, ISs, and SURs that are  used in the
      extraction batch. The LRB is used to determine if the method analyte or other
      interferences are present in the laboratory environment, the reagents, or the apparatus.

3.12  LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The
      single-laboratory LCMRL is the lowest true concentration for which the future
      recovery is predicted to fall, with high confidence (99%), between 50 and 150%
      recovery.1

3.13  MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by
      vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and
      reactivity data including storage, spill, and handling precautions.

3.14  MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
      reported by a laboratory as a quantitated value for a target analyte in a sample
      following analysis. This defined concentration must meet the criteria defined in
      Section 9.2.4 and must not be any lower than the concentration of the lowest
      calibration standard for that analyte.  A laboratory may be required to demonstrate a
      specific MRL by a regulatory body if this method is being performed for compliance
      purposes.

3.15  QUALITY CONTROL SAMPLE (QCS) - A sample or standard prepared using a
      Stock Standard Solution (SSS) of the method analyte that is obtained from a source
      external to the laboratory and different from the source of calibration standards.  The
      second  source SSS is used to fortify the QCS at a known concentration.  The QCS is
      used to  check calibration standard integrity.

3.16  SELECTED ION MONITORING (SIM) - A GC/MS technique where only one or a
      few ions are monitored. When used with gas chromatography, the set of ions
      monitored is usually changed periodically throughout the chromatographic run, to
      correlate with the characteristic ions of the analyte,  SUR and IS as they elute from the
      chromatographic column.

3.17  STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing the
      method analyte prepared in the laboratory using assayed reference materials or
      purchased from a reputable commercial source.
                                    522-5

-------
   3.18  SURROGATE ANALYTE (SUR) - A pure analyte, which is extremely unlikely to be
         found in any sample, and which is added to a sample aliquot in a known amount
         before extraction or other processing, and is measured with the same procedures used
         to measure other sample components.  The purpose of the SUR is to monitor method
         performance with each sample.

4.  INTERFERENCES

   4.1   All glassware must be meticulously cleaned.  Wash glassware with detergent and tap
         water, rinse with tap water, followed by reagent water.  Non-volumetric glassware
         should be heated in a muffle furnace at 400 °C for 2 hours. Volumetric glassware
         should be solvent rinsed with DCM or purge and trap grade methanol after washing,
         and dried in a low temperature oven (<120 °C) or air dried.

   4.2   Method interferences may be caused by contaminants in solvents, reagents (including
         reagent water), sample bottles and caps, and other sample processing hardware that
         lead to discrete artifacts and/or elevated baselines in the chromatograms. All items
         such as these must be routinely demonstrated to be free from interferences (less than
         */3 the MRL for the method analyte) under the conditions of the analysis by analyzing
         LRBs as described in Sections 9.2.1 and 9.3.1.  Subtracting  blank values from
         sample  results is not permitted.

   4.3   Purge and trap grade methanol must be used for all steps where methanol is used
         in this method.  Other grades of methanol contain numerous low molecular
         weight compounds that contain interfering ions which may prohibit accurate
         identification and  quantitation of the analyte, SUR and IS.

   4.4   Matrix interferences may be caused by contaminants that are co-extracted from the
         sample.  The extent of matrix interferences will vary considerably from source to
         source, depending upon the nature of the water.

   4.5   Preservatives (Sect. 8.1) are added to samples to ensure sample stability during
         shipping and storage prior to analysis.  The potential exists for trace-level organic
         contaminants in these reagents. Interferences from these sources should be monitored
         by analysis of LRBs, particularly when new lots of reagents are acquired.

   4.6   Solid phase extraction cartridges may be a source of interferences.  The analysis of
         field and laboratory reagent blanks can provide important information regarding the
         presence or absence of such interferences.  Brands and lots of SPE devices should be
         tested to ensure that contamination does not preclude analyte identification and
         quantitation.

   4.7   Analyte carry-over may occur when a relatively "clean" sample is analyzed
         immediately after a sample containing relatively high concentrations of compounds.
         Syringes and splitless injection port liners must be cleaned carefully or replaced as
         needed.  After analysis of a sample containing high concentrations of compounds, a

                                         522-6

-------
         LRB should be analyzed to ensure that accurate values are obtained for the next
         sample. During automated GC/MS analyses, extracts with positive results that were
         analyzed immediately following a sample with high concentrations of the analyte,
         should be reanalyzed after analyzing an acceptable LRB.  If the analyte is not detected
         in extracts analyzed immediately after a high concentration extract, no reanalysis is
         necessary.

   4.8   Silicone compounds may be leached from punctured autosampler vial septa,
         particularly when particles of the septa sit in the vial. This can occur after repeated
         injections from the same autosampler vial.  If this method is performed in full scan
         mode, these silicone compounds may appear as regularly spaced chromatographic
         peaks with similar fragmentation patterns.  They can unnecessarily complicate the
         total ion chromatogram and may cause interferences at high levels.

   4.9   High laboratory background levels of 1,4-dioxane have been reported to be associated
         with air contamination. If 1,4-dioxane is detected in LRBs, room air should be
         considered as a possible source.3'4

5.  SAFETY

   5.1   The toxicity or carcinogenicity of each reagent used in this method has not been
         precisely defined. Each chemical should be treated as a potential health hazard, and
         exposure to these chemicals should be minimized.  1,4-Dioxane is classified as a class
         B2 or probable human carcinogen.  Each laboratory is responsible for maintaining an
         awareness of OSHA regulations regarding safe handling of chemicals used in this
         method.  A reference file  of MSDSs should be made available to all personnel
         involved in the chemical analysis. Additional references on laboratory safety are
         available.6"8

   5.2   Pure standard materials and stock standard  solutions of these compounds should be
         handled with suitable protection to skin and eyes, and care should be taken not to
         breathe the vapors or ingest the materials.

   5.3   Sodium bisulfate is used as a sample preservative to inhibit microbial growth  and
         potential decay of 1,4-dioxane. Sodium bisulfate is highly acidic and should be used
         with appropriate caution.

6.  EQUIPMENT AND SUPPLIES (References to  specific brands or catalog numbers are
   included for illustration only, and do not imply endorsement of the product.)

   6.1   SAMPLE CONTAINERS - Glass bottles fitted with polytetrafluoroethylene (PTFE)
         lined screw caps.  Although samples do not need to be collected headspace free, the
         size of the sample bottle should be selected based upon the sample volume to  be
         extracted.
                                         522-7

-------
6.2   VIALS - Various sizes of glass vials with PTFE-lined screw caps for storing standard
      solutions and extracts, including glass 2-mL autosampler vials with PTFE-faced septa.

6.3   VOLUMETRIC FLASKS - Class A, various sizes, including 1, 5, and 10 mL for
      preparation of standards. A 2-mL volumetric collection tube may also be used for the
      collection of the extract when SPE Option 2 is used (Sect. 11.5.3).

6.4   GRADUATED CYLINDERS - Glass, various sizes, including 100 and 500 mL for
      measurement of sample volumes.

6.5   MICRO  SYRINGES - Suggested sizes include 10, 25,  50, 100, 250, 500, and
      1000 uL.

6.6   CONICAL CENTRIFUGE TUBES - 15 mL, glass, with graduations at 0.1 mL and
      PTFE-lined screw caps (KEVIAX #45166 or equivalent), suitable for collection of the
      eluent from the 2-g solid phase cartridge after extraction.

6.7   ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g.

6.8   SOLID PHASE EXTRACTION (SPE) APPARATUS

   6.8.1   SPE CARTRIDGES - SPE carbon cartridges acceptable for use in this method
          are made of either coconut charcoal or  synthetic carbon as described in Sections
          6.8.1.1 and 6.8.1.2 below. Graphitized carbon does not retain 1,4-dioxane and
          may not be used in this method.

       6.8.1.1   Option 1 (for use with sample volumes of < 500 mL) - 6-mL polypropylene
               tubes packed with 2 g coconut charcoal (80-120 mesh, approximately
               150 um).  At the time of method development, pre-packed cartridges were
               obtained from Restek Corporation (cat. # 26032) and United Chemical
               Technologies (UCT; cat. # EU52112M6).  Equivalent brands of coconut
               charcoal SPE cartridges with the  specified  particle size and whose
               performance meets all QC criteria in Section 9 may be used.

       6.8.1.2   Option 2 (for use with samples volumes of < 100 mL) - Waters AC-2 Sep-
               Pak 400-mg activated carbon SPE cartridge (cat. # JJAN20229; 85 um
               particle size) or Supelco Supelclean ENVI-Carb Plus 400-mg synthetic
               carbon SPE cartridge (cat. # 54812-U).

   6.8.2   VACUUM EXTRACTION MANIFOLD - Equipped with flow/vacuum control
          (Supelco cat. # 57250-U or equivalent).

   6.8.3   SAMPLE DELIVERY SYSTEM - Use of a transfer tube system (Supelco
          "Visiprep", cat. # 57275 or equivalent), which transfers the sample directly from
          the sample container to the SPE cartridge, is recommended. Sample reservoirs
                                    522-8

-------
          (Varian, cat. #12131012 or equivalent), which attach to the cartridge, may be
          used, although they hold only a limited volume of sample.

   6.8.4   An automatic or robotic system designed for use with SPE cartridges may be used
          if all QC requirements discussed in Section 9 are met.  Automated systems may
          use either vacuum or positive pressure to process samples and solvents through
          the cartridge  All extraction and elution steps must be the same as in the
          manual procedure. Extraction and/or elution steps may not be changed or
          omitted to accommodate the use of an automated system.

6.9   LABORATORY OR ASPIRATOR VACUUM SYSTEM - Sufficient capacity to
      maintain a vacuum of approximately 15 to 25 inches of mercury.

6.10  GAS CHROMATOGRAPH/MASS SPECTROMETER (GC/MS) SYSTEM

   6.10.1  FUSED SILICA CAPILLARY GC COLUMN - Data presented in this method
          were obtained with a column typically used for volatiles analysis: Varian CP
          Select 624, 30 m x 0.25-mm i.d. with a 1.4 um film thickness, or equivalent.
          Other types of columns may be used. However, use of shorter or thinner film
          columns is not recommended because this will result in loss of resolution and
          very short retention times.  Because the method analyte, SUR and IS have only
          very low mass ions in their spectra, and SIM is being used, it is important to strive
          for chromatographic separation from any potential interferences. See Section 9.4
          regarding the use of alternate GC  columns.

          CAUTION: If a GC column and/or temperature program different than that
          described in this method is used, the analyst must verify that the temperature
          program and subsequent column bake-out is sufficient to remove all injected
          material from the column. Although the retention times of the method analyte,
          IS and SUR are short, many chemicals may be co-extracted by carbon SPE and
          these need to be eluted from the GC column to prevent them from appearing as
          interferences in subsequent analyses.

   6.10.2  GC INJECTOR AND OVEN - The performance data in Section 17 were obtained
          using hot, splitless injection. Other injection techniques such as temperature
          programmed injections, cold on-column injections and large volume injections
          may be used if the QC criteria in Section 9 are met.  Equipment designed
          appropriately for these alternate types of injections must be used if these options
          are selected.

   6.10.3  GC/MS INTERFACE - The interface should allow the capillary column or
          transfer line exit to be placed within a few millimeters of the ion source.  Other
          interfaces, such as jet separators, are acceptable as long as the system has
          adequate sensitivity and QC performance criteria in  Section 9 are met.
                                     522-9

-------
       6.10.4 MASS SPECTROMETER (MS) - Any type of mass spectrometer may be used
             (i.e., quadrupole, ion trap, time of flight, etc.), although the SIM option may not
             provide enhanced sensitivity, or be an available option on all instruments. The
             spectrometer must produce a mass spectrum that meets all criteria in Table 1
             when a solution containing approximately one to two nanograms of bromo-
             fluorobenzene (BFB) is injected into the GC/MS. This test must be performed in
             the full scan mode.  Use a single spectrum at the apex of the BFB peak, an
             average spectrum of the three highest points of the peak, or an average spectrum
             across the entire peak to evaluate the performance of the system.  Appropriate
             background subtraction is permitted. The scan time should be set so that there is a
             minimum of five scans across the chromatographic peak. Ten scans across
             chromatographic peaks are recommended.

       6.10.5 DATA SYSTEM - An interfaced data system is required to acquire, store, and
             output MS data. The computer software should have the capability of processing
             stored GC/MS data by recognizing a GC peak within a given retention time
             window.  The software must allow integration of the ion abundance of any
             specific ion between specified times or scan number limits. The software must be
             able to construct linear regressions and quadratic calibration curves, and calculate
             analyte concentrations.

   6.11  N, N-DIETHYL-P-PHENYLENEDIAMINE (DPD) CHLORINE TEST KIT - Used
         to verify sample dechlorination when samples are received at the analytical laboratory
         (Hach model CN-66; cat. # 2231-01 or equivalent).

7.  REAGENTS AND STANDARDS

   7.1   REAGENTS AND SOLVENTS - Reagent grade or better chemicals should be used in
         all steps.  Unless otherwise indicated, it is intended that all reagents will conform to
         the specifications of the Committee on Analytical Reagents of the American Chemical
         Society (ACS), where such specifications are available.  Other grades may be used,
         provided it is first determined that the reagent is of sufficiently high purity to permit its
         use without lessening the quality of the determination. During  method development,
         only purge and trap  grade methanol was found to be sufficiently free from  low
         molecular weight interferences (as described in Sect. 4.3). Therefore that grade must
         be used.

       7.1.1  HELIUM - 99.999 % or better, GC carrier gas.

       7.1.2  REAGENT WATER - Purified water which does not contain any measurable
             quantities of the method analyte or interfering compounds at or above l/3 the
             MRL for 1,4-dioxane.  In addition, there must be no interferences with the SUR
             or IS.

       713  METHANOL (CASRN 67-56-1) - Purge and trap grade only (see Section 4.3V
                                       522-10

-------
   7.1.4   DICHLOROMETHANE (DCM) (CASRN 75-09-02) - High purity, demonstrated
          to be free of analytes and interferences (B&J Brand GC2, Capillary GC or GC/MS
          Grade or equivalent).

   7.1.5   SODIUM SULFATE, ANHYDROUS (CASRN 7757-82-6) - Soxhlet extracted
          with DCM for a minimum of four hours or heated to 400 °C for two hours in a
          muffle furnace. An "ACS grade, suitable for pesticide residue analysis", or
          equivalent, of anhydrous sodium sulfate is recommended.

   7.1.6   SAMPLE PRESERVATION REAGENTS

      7.1.6.1   SODIUM SULFITE (CASRN 7757-83-7) - Reduces free and combined
               chlorine in samples that have been disinfected with chlorine and/or
               chloramine.

      7.1.6.2   SODIUM BISULFATE (CASRN 7681-38-1) - Anhydrous, technical grade.
               It is added to acidify the samples to pH < 4 to act as a microbial inhibitor
               during sample shipping and storage.

7.2   STANDARD SOLUTIONS - When a compound purity is assayed to be 96% or
      greater, the weight can be used without correction to calculate the concentration of the
      stock standard. Solution concentrations listed in this section were used to develop this
      method and are included as examples. Alternate concentrations may be used as
      necessary depending on instrument sensitivity  and the calibration range used.
      Standards for sample fortification generally should be prepared in the smallest volume
      that can be accurately measured to minimize the addition of excess organic solvent to
      aqueous samples. During method development, standard solutions were stored in a
      freezer at about -15 to -20 °C for several months. Experience indicates that the most
      likely cause of standard deterioration for this method is solvent evaporation.
      Therefore, it is recommended that standard solutions be stored at 0 °C or less, with
      minimal headspace. Laboratories should use standard QC practices to determine when
      standards need to be replaced. Standard QC practices, such as monitoring area counts
      and recovery percentages of CCCs, would help to determine if a standard requires
      replacement.  As mentioned above, the major form of "standard deterioration" is
      solvent evaporation, which would result in a continuous increase in area counts, and in
      some cases, target and SUR recoveries. During method development, SSSs and
      dilutions of SSSs of 1,4-dioxane, THF-Jg, 1,4-dioxane-dg, and BFB were valid for at
      least 6 months after opening/preparation.  Expiration dates provided by the vendor
      should be used as an indicator of SSS shelf life prior to opening.

      NOTE: Only purge and trap grade methanol may be used in the preparation of
      standards and sample fortification solutions.

   7.2.1   STOCK STANDARD SOLUTIONS (SSS) - Stock standard solutions of the
          method analyte, SUR and IS are commercially available from multiple
          commercial sources. It is recommended that these solutions be purchased from

                                    522-11

-------
       commercial sources. However, they may be prepared from neat materials.
       Utilizing stock standards prepared in methanol allows the same stock standard to
       be used to prepare both calibration standards and sample fortification solutions.
       The concentrations of the stock standard solutions used during method
       development were 1000-2000 ug/mL.

   7.2.1.1  PREPARATION OF STOCK STANDARD SOLUTIONS (SSS) FROM
           NEAT MATERIALS

       7.2.1.1.1  Place about 9.8 mL of methanol in a 10-mL volumetric flask. Allow
                the flask to stand, unstoppered, for about 10 min or until all alcohol
                wetted surfaces have dried. Weigh to the nearest 0.1 mg.

       7.2.1.1.2  Using a 100 uL-syringe, quickly add two or more drops of the neat
                standard material to the flask. Be sure that the standard falls directly
                into the alcohol, without contacting the neck of the flask.

       7.2.1.1.3  Quickly reweigh, dilute to volume and mix by inverting the flask
                several times. Calculate the concentration in  ug/uL from the net gain
                in weight.

       7.2.1.1.4  Store SSS in 12-15 mL glass vials with PTFE lined screw caps.

7.2.2   CALIBRATION STANDARD SOLUTIONS (CAL) - Prepare a series of
       calibration standards to encompass the desired calibration range.  Calibration
       standards must contain varying amounts  of 1,4-dioxane,  and a fixed amount of
       both the SUR and the IS, and be prepared in DCM.  The  number of standards
       required is determined by the calibration range.  Three standards are required for
       one order of magnitude, six standards for two orders of magnitude and nine
       standards for three orders of magnitude.  The calibration curve associated with the
       demonstration data in Section 17 contained nine standards with concentrations of
       1,4-dioxane ranging from 0.002 ug/mL to 1.0 ug/mL. (This corresponds to a
       concentration range of 0.04-20 ug/L in the drinking water samples.)  The
       concentrations of both l,4-dioxane-t/8 (SUR) and the THF-J8 (IS) were
       0.500 ug/mL in each standard.

7.2.3   SAMPLE FORTIFICATION SOLUTIONS

   7.2.3.1  ANALYTE FORTIFICATON SOLUTION - Prepare one or more solutions
           in methanol for use in preparing LFBs and LFSMs  (Sect. 11.3.5).  The
           number of solutions needed depends upon the calibration range and/or
           sample volume.  During method development, two solutions at
           concentrations of 20 ug/mL and 200 ug/mL were used to fortify 100-mL
           and 500-mL samples, respectively.
                                 522-12

-------
          7.2.3 .2  SURROGATE ANALYTE (SUR) FORTIFICATION SOLUTION -
                  Prepare one or more solutions of l,4-dioxane-6?8 in methanol to be used for
                  fortification of the SUR into samples, LFBs, LRBs and LFSMs
                  (Sect. 11.3.4).  The concentrations used during method development were
                  200 ug/mL and 2000 ug/mL (STOCK) to fortify 100-mL and 500-mL
                  samples, respectively.

          7.2.3 .3  INTERNAL STANDARD (IS) FORTIFICATION SOLUTION - Prepare
                  one or more solutions of THF-Jg in DCM to be used to add the IS to all
                  extracts (Sect. 11.4.4 and 11.5.4). During method development, the
                  concentrations of these solutions were 100 ug/mL and 1000 ug/mL
                  (STOCK), used to fortify 2-mL and 10-mL extracts, respectively.

       7.2.4  GC/MS TUNE CHECK SOLUTION - Stock standard solutions of BFB (CASRN
             460-00-4) are available commercially. Prepare a BFB solution at a concentration
             of 1-2 ug/mL in DCM by dilution of the stock standard. Store this solution in an
             amber glass screw cap vial at 0 °C or less. To prepare a SSS of BFB from neat
             material, see Section 7.2.1.1.

8.  SAMPLE COLLECTION, PRESERVATION, AND STORAGE
8.1
         SAMPLE PRESERVATIVES - Preservation reagents, listed in the table below, are
         added to each sample at the time of sample collection. Sodium sulfite must be added
         first and may be placed as a dry material in the sample bottles prior to shipment to the
         field.  Aqueous solutions of sodium sulfite may not be added to sample bottles prior to
         shipment to the collection site because these solutions are unstable and cannot be
         relied upon to completely dechlorinate the samples.  Sodium bisulfate is added only
         after the sodium sulfite has been dissolved in the aqueous sample. See Section 8.2.2
         and 8.2.3 for complete instructions.
Compound
Sodium sulfite
Sodium bisulfate
Amount
50mg/L
1 g/L (approx)
Purpose
Reduce chlorine/chloramine residual
Microbial inhibitor
   8.2   SAMPLE COLLECTION

       8.2.1   Open the tap and allow the system to flush until the water temperature has
             stabilized (approximately three to five min). Collect samples from the flowing
             system.

       8.2.2   Fill sample bottles, taking care not to flush out the sample dechlorination reagent.
             Samples do not need to be collected headspace free.

       8.2.3   After collecting the sample, cap the bottle and agitate by hand until the sodium
             sulfite is dissolved.  Add enough sodium bisulfate such that the final
                                       522-13

-------
             concentration will be 1 g/L. Cap the bottle and mix until dissolved. Unless field
             verification of pH is to be performed, keep the sample sealed until just prior to
             extraction.

       8.2.4  Field verification of pH 4 (optional).  It is anticipated that 1 g/L of sodium
             bisulfate will be sufficient to acidify most samples to < pH 4.  If there is reason to
             suspect that more may be needed, the pH can be verified with narrow range pH
             paper at the time of sample collection. After acidification and mixing, pour a
             small amount of sample over a strip of the pH paper (do not dip the strip in the
             sample). Read the result as instructed on the pH paper package. If the pH is > 4,
             add additional sodium bisulfate until pH < 4 is obtained.  Seal the bottle, and keep
             the sample sealed until extraction.

   8.3   SAMPLE SHIPMENT AND STORAGE - Samples must be chilled during shipment
         and must not exceed 10 °C during the first 48 hours after collection.  Sample
         temperature must be confirmed to be at or below 10 °C when they are received at the
         laboratory. Depending upon the water temperature at the time of collection, samples
         may need to be refrigerated to reduce their temperature before being packed on ice for
         shipment. Samples stored in the lab must be held at or below 6 °C until extraction,  but
         should not be frozen.  Freezing samples may compromise the sealed cap or result in
         sample bottle breakage.

       8.3.1  Verification of sample dechlorination - Upon the receipt of samples at the
             laboratory, verify that Field Samples were dechlorinated at the time of collection.
             The absence of total chlorine can be verified with a DPD chlorine test kit (Sect.
             6.11).

   8.4   SAMPLE AND EXTRACT HOLDING TIMES - Aqueous samples may be stored  as
         described above for up to 28 days from collection. Sample extracts, as prepared in
         Section 11, may be stored at -5 °C and protected from light for an additional 28 days.

9.  QUALITY CONTROL

   9.1   QC requirements include the Initial Demonstration of Capability (IDC) and ongoing
         QC requirements that must be met when preparing and analyzing Field Samples.  This
         section describes each QC parameter, its required frequency, and the performance
         criteria that must be met in order to meet USEPA quality objectives. The QC criteria
         discussed in the following sections are summarized in Section 17, Tables 6 and 7.
         These QC requirements are considered the minimum acceptable QC criteria.
         Laboratories are encouraged to institute additional QC practices to meet their specific
         needs.  If method modifications are under consideration by the laboratory, refer to
         Sect. 9.4  for detailed requirements.

   9.2   INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be
         successfully performed prior to analyzing any Field Samples. Prior to conducting the
                                        522-14

-------
  IDC, the analyst must first generate an acceptable Initial Calibration following the
  procedure outlined in Section 10.2.

9.2.1   INITIAL DEMONSTRATION OF LOW BACKGROUND - Before any samples
       are analyzed, or any time a new lot or brand of SPE materials or solvents are
       received from a supplier, it must be demonstrated that a LRB is reasonably free of
       any contamination that would prevent the identification or quantitation of
       1,4-dioxane, the IS or SUR.

   9.2.1.1   A source of potential contamination is the SPE materials which may contain
            phthalate esters, silicon compounds, and other contaminants that could
            interfere with the determination of the method analyte, SUR or IS.
            Although extraction media are generally made of inert materials, they may
            still contain extractable organic material.  If the background contamination
            is sufficient to prevent accurate and precise measurements, the condition
            must be corrected before proceeding with the IDP (Sect. 9.2.2).

   9.2.1.2   Other sources of background contamination are solvents, reagents (including
            reagent water), room air and glassware. Background contamination must be
            reduced to an acceptable level before proceeding with the IDP (Sect. 9.2.2).
            Background from 1,4-dioxane and interferences must be < 1/3 the MRL. If
            background contamination is an on-going or intermittent issue, the analyst
            must not attempt to verify an MRL less than the mean LRB concentration
            + 3o, or three times the mean LRB concentration, whichever is greater.

            NOTE: Although quantitative data below the MRL may not be reliably
            accurate enough for data reporting, such data are useful in determining an
            MRL cut off for the method analyte if it is found in the LRB.  Therefore,
            blank contamination levels may be estimated by extrapolation, when the
            concentration is below the lowest calibration standard.  This also applies to
            estimating LRB concentrations for daily low background verification.

9.2.2   INITIAL DEMONSTRATION OF PRECISION (IDP) - Prepare, extract, and
       analyze four to seven replicate LFBs fortified near the midrange of the initial
       calibration curve according to the procedure described in Section 11. Sample
       preservatives as described in Section 8.1 must be added to these samples.  The
       relative standard deviation (RSD) of the results of the replicate analyses must be
       less than 20%.

9.2.3   INITIAL DEMONSTRATION OF ACCURACY (IDA) - Using the same set of
       replicate data generated for Section 9.2.2, calculate average recovery. The
       average recovery of the replicate values must be within ± 20% of the true value.

9.2.4   MINIMUM REPORTING LEVEL (MRL)  CONFIRMATION - Establish a target
       concentration for the MRL for  1,4-dioxane  based on the intended use of the
       method.  In some cases, the MRL may be dictated by USEPA or another

                                 522-15

-------
       regulatory body. Establish an Initial Calibration following the procedure outlined
       in Section 10.2.  The lowest calibration standard used to establish the Initial
       Calibration (as well as the low-level CCC) must be at or below the concentration
       of the MRL.  Establishing the MRL concentration too low may cause repeated
       failure of ongoing QC requirements. Validate the MRL following the procedure
       outlined below.

   9.2.4.1   Fortify, extract, and analyze seven replicate LFBs at the proposed MRL
            concentration. These LFBs must contain all method preservatives described
            in Section 8.1.  Calculate the mean (Mean) and standard deviation for these
            replicates.  Determine the Half Range for the prediction interval of results
            (HRpiR) using the equation below

                                  HRpm =  3.963S

            where S is the standard deviation, and 3.963  is the constant value for seven
            replicates.1

            NOTE: The mass spectrum (either SIM or full scan) for the method analyte
            in the LFBs must meet  all the analyte identification criteria in Sections 12.1
            and 12.2, i.e., the MRL verification may not be performed on LFBs where
            only the base peak is observed. If during MRL confirmation all
            identification ions are not observed, the MRL selected is too low.

   9.2.4.2   Confirm that the upper  and lower limits for the Prediction Interval of Result
            (PIR = Mean +_ HRpjR) meet the upper and lower recovery limits as shown
            below.

                The Upper PIR Limit must be < 150%  recovery.

                              Mean + HRpm
                         	^	xlOO% <150%
                         FortifiedConcentration

                The Lower PIR Limit must be > 50% recovery.

                               Mean-HRpm
                          	^	xlOO% > 50%
                          FortifiedConcentration

   9.2.4.3   The MRL is validated if both the Upper and  Lower PIR Limits meet the
            criteria described above (Sects. 9.2.4.2). If these criteria are not met, the
            MRL for 1,4-dioxane has been set too low and must be re-evaluated at a
            higher concentration.

9.2.5   CALIBRATION CONFIRMATION - Analyze a QCS as described in Section
       9.3.9 to confirm the accuracy of the standards and calibration curve.

                                  522-16

-------
   9.2.6  DETECTION LIMIT (DL) DETERMINATION (optional) - While DL
          determination is not a specific requirement of this method, it may be required by
          various regulatory bodies associated with compliance monitoring. It is the
          responsibility of the laboratory to determine ifDL determination is required
          based upon the intended use of the data.

          Replicate analyses for this procedure should be done over at least three days (i.e.,
          both the sample extraction and the GC analyses should be done over at least three
          days).  Prepare at least seven replicate LFBs containing 1,4-dioxane at a
          concentration estimated to be near the DL. This concentration may be estimated
          by selecting a concentration at two to five times the noise level.  The appropriate
          fortification concentrations will be dependent upon the sensitivity of the GC/MS
          system used. All preservation reagents listed in Section 8.1 must also be added to
          these samples.  Analyze the seven replicates through all steps of Section 11.

          NOTE: If an MRL confirmation data set meets these requirements, a DL may be
          calculated from the MRL  confirmation data, and no additional analyses are
          necessary.

          Calculate the DL using the equation

                  l—sl—i — *J <"* / /   -11     r\ r\r\\
                           (n-\ l-ctr=0.99)

          where:
                 s = standard deviation of replicate analyses
                 t («-i, i-a=o.99) = Student's t value for the 99% confidence level with n-\
                  degrees of freedom
                 n = number of replicates.

          NOTE: There are no precision and accuracy requirements for the DL replicates.
          Do not subtract blank values when performing DL calculations.

9.3    ONGOING QC REQUIREMENTS - This section summarizes the ongoing QC
      criteria that must be followed when processing and analyzing  Field Samples.

   9.3.1  LABORATORY REAGENT BLANK (LRB) - An LRB  is required with each
          extraction batch to confirm that potential background contaminants are not
          interfering with the identification or quantitation of the method analyte, SUR or
          IS.  If the LRB produces a peak within the retention time  window of any analyte
          that would prevent the identification or quantitation of that analyte, determine the
          source of contamination and eliminate the interference before processing samples.
          Background contamination must be reduced to an acceptable level before
          proceeding.
                                     522-17

-------
       NOTE: Absence of interferences must be verified for 1,4-dioxane at both the
       quantitation ion (QI) (m/z 88) and the confirmation ion (m/z 58). Background
       from contaminants that interfere with the measurement of 1,4-dioxane must be
       below */3 of the MRL. Blank contamination may be estimated by extrapolation, if
       the concentration is below the lowest calibration standard.  This procedure is not
       allowed for sample results as it may not meet data quality objectives. If the
       method analyte is detected in the LRB at concentrations equal to or greater than
       this level, then all data for the analyte must be considered invalid for all samples
       in the extraction batch.

9.3.2   CONTINUING CALIBRATION CHECK (CCC) - CCC Standards are analyzed
       at the beginning of each analysis batch, after every ten Field Samples, and at the
       end of the analysis batch. See Section 10.3 for concentration requirements and
       acceptance criteria.

9.3.3   LABORATORY FORTIFIED BLANK (LFB) - An LFB is required with each
       extraction batch. The fortified concentration of the LFB must be rotated between
       low, medium, and high concentrations from batch to batch. The low
       concentration LFB must be as near as practical to, but no more than two times the
       MRL.  Similarly, the high concentration LFB should be near the high end of the
       calibration range established during the initial  calibration (Sect. 10.2). Results of
       the low-level LFB analyses must be 50-150%  of the true value. Results of the
       medium and high-level LFB analyses must be 70-130% of the true value.  If the
       LFB results do not meet these criteria for  1,4-dioxane, then all data must be
       considered invalid for all samples in the extraction batch.

9.3.4   MS TUNE CHECK - A complete description  of the MS Tune Check is found in
       Section 10.2.1. Acceptance criteria for the MS Tune Check are summarized in
       Section 17, Table 1.  The MS Tune Check must be performed each time a major
       change is made to the mass spectrometer,  and  prior to establishing and/or re-
       establishing an initial calibration (Sect. 10.2).  In this method, daily BFB analysis
       is not required.

9.3.5   INTERNAL STANDARDS (IS) - The analyst must monitor the peak area of the
       IS in all injections during each analysis day. The IS response (peak area) in any
       chromatographic run must not deviate from the response in the most recent CCC
       by more than 30%, and must not deviate by more than 50% from the mean area
       measured during initial analyte calibration. If the IS area in a chromatographic
       run  does not meet these criteria, inject a second aliquot of that extract.

   9.3.5.1   If the reinjected aliquot produces an acceptable IS response, report results
            for that aliquot.

   9.3.5.2   If the reinjected extract fails again, the analyst should check the calibration
            by reanalyzing the most recently acceptable calibration standard.  If the
            calibration standard fails the criteria of Section 10.3, recalibration is in order

                                  522-18

-------
            per Section 10.2. If the calibration standard is acceptable, extraction of the
            sample may need to be repeated provided the sample is still within the
            holding time.  Otherwise, report results obtained from the reinjected extract,
            but annotate as "suspect/IS recovery." Alternatively, collect a new sample
            and re-analyze.

9.3.6   SURROGATE (SUR) RECOVERY - The SUR standard is fortified into the
       aqueous portion of all samples, LRBs, LFBs, CCCs, LFSMs and LFSMDs prior
       to extraction.  It is also added to the calibration standards.  The SUR is a means of
       assessing method performance from extraction to final chromatographic
       measurement. Calculate the percent recovery (%R) for the SUR using the
       equation

                                       ( A\
                                 %R =  — xlOO
                                       UJ

       where:
             A = calculated SUR concentration for the QC or Field Sample
             B = fortified concentration of the SUR.

   9.3.6.1   SUR recovery must be in the range of 70-130% of the true value.  When
            SUR recovery is less than 70% or greater than 130%, check the 1)
            calculations to locate possible errors, 2) standard solutions for degradation,
            3)  contamination, and 4) instrument performance. Correct the problem and
            reanalyze the  extract.

   9.3.6.2   If the extract reanalysis meets the SUR recovery criterion, report only data
            for the reanalyzed extract.

   9.3.6.3   If the extract reanalysis fails the 70-130% recovery criterion, the analyst
            should check the calibration by injecting the last CCC that passed. If the
            CCC fails the criteria of Section 9.3.6.1, recalibration is in order per Section
            10.2. If the calibration standard is acceptable, extraction of the sample
            should be repeated provided the sample is still within the holding time. If
            the re-extracted sample also fails the recovery criterion, report all data for
            that sample as "suspect/SUR recovery" to inform the data user that the
            results are suspect due to SUR recovery.  Alternatively, collect a new
            sample and re-analyze.

9.3.7   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of an
       LFSM is required in each extraction batch and is used to determine that the
       sample matrix does not adversely affect method accuracy. If the occurrence of
       1,4-dioxane in the samples is infrequent, or if historical trends are unavailable, a
       LFSMD must be prepared, extracted, and analyzed from a duplicate Field Sample
       to assess method precision (Sect. 9.3.8). Extraction batches that contain LFSMDs
       do not require the analysis of a Field Duplicate. If a variety of different sample

                                  522-19

-------
       matrices are analyzed regularly, for example, drinking water from groundwater
       and surface water sources, method performance should be established for each.
       Over time, LFSM data should be documented by the laboratory for all routine
       sample sources analyzed.

   9.3.7. 1   Within each extraction batch, a minimum of one Field Sample is fortified as
            an LFSM for every 20 samples extracted. The LFSM is prepared by spiking
            a sample with an appropriate amount of the Analyte Fortification Solution
            (Sect. 7.2.3.1). Select a spiking concentration that is greater than or equal to
            the matrix background concentration, if known.  Use historical  data and
            rotate through low, medium and high concentrations when selecting a
            fortifying concentration.

   9.3.7.2   Calculate the percent recovery (%R) for the analyte using the equation
                                             C

            where:
                   A = measured concentration in the fortified sample
                   B = measured concentration in the unfortified sample
                   C = fortification concentration.

            NOTE: Field Samples that have native analyte concentrations above the
            DL but below the MRL, and are fortified at concentrations at or near the
            MRL, should be corrected for the native levels in order to obtain
            meaningful %R values. This example and the LRB (Sect. 9.3.1) are the only
            permitted use of analyte results below the MRL.

   9.3.7.3   Analyte recoveries may exhibit matrix bias. For samples fortified at or
            above their native concentration, recoveries should range between 70-130%,
            except for low-level fortification near or at the MRL (within a factor of two
            times the MRL concentration) where 50-150% recoveries are acceptable. If
            the accuracy  of any analyte falls outside the designated range, and the
            laboratory performance for that analyte  is shown to be in control in the
            CCCs, the recovery is judged to be matrix biased. The result for that analyte
            in the unfortified sample is labeled "suspect/matrix" to inform the data user
            that the results are suspect due to matrix effects.

9.3.8   FIELD DUPLICATE (FD) OR LABORATORY FORTIFIED SAMPLE
       MATRIX DUPLICATE (LFSMD) - Within each extraction batch, a minimum of
       one Field Duplicate (FD) or Laboratory Fortified Sample Matrix Duplicate
       (LFSMD) must be analyzed. Duplicates check the precision associated with
       sample collection,  preservation, storage,  and laboratory procedures. If method
       analytes are not routinely observed in Field Samples, an LFSMD  should be
       analyzed rather than an FD.

                                 522-20

-------
   9.3.8.1   Calculate the relative percent difference (RPD) for duplicate FD
            measurements (FD1 and FD2) using the equation

                                FDI-FD2
                        RPD=-,	:—xlOO
                              (FD\ + FD2}/2

   9.3.8.2   RPDs for FDs should be < 30%. Greater variability may be observed when
            FDs have analyte concentrations that are within a factor of two of the MRL.
            At these concentrations, FDs should have RPDs that are < 50%. If the RPD
            of the analyte falls outside the designated range, and the laboratory
            performance for the analyte is shown to be in control in the CCC, the
            recovery is judged to be matrix biased. The result for that analyte in the
            unfortified sample is labeled "suspect/matrix" to inform the data user that
            the results are suspect due to matrix effects.  Failure  to meet the FD criteria
            is a reflection of the performance of the method for that individual sample,
            and does not constitute analysis or extraction batch failure.

   9.3.8.3   If an LFSMD is analyzed instead of a FD, calculate the relative percent
            difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the
            equation

                              LFSM-LFSMD
                    RPD=-	:—XlOO
                            (LFSM+LFSMD} 12

   9.3.8.4   RPDs for duplicate LFSMs should be < 30% for samples fortified at or
            above their native concentration.  Greater variability may be observed when
            LFSMs are fortified at analyte concentrations that are within a factor of two
            of the MRL.  LFSMs fortified at these concentrations should have RPDs
            that are < 50% for samples fortified at or above their native concentration.
            If the RPD of the analyte falls outside the designated range, and the
            laboratory performance for the analyte is shown to be in control in the CCC,
            the recovery is judged to be matrix biased. The result for that analyte in the
            unfortified sample is labeled "suspect/matrix" to inform the data user that
            the results are suspect due to matrix effects. Failure to meet the
            LFSM/LFSMD criteria is a reflection of the performance of the method for
            that individual sample, and does not constitute analysis or extraction batch
            failure.

9.3.9   QUALITY CONTROL SAMPLES (QCS) - As part of the IDC (Sect. 9.2), each
       time new calibration standards are prepared, and at least quarterly, analyze a QCS
       sample from a source different from the source of the calibration standards. The
       QCS  should be prepared and analyzed just like a CCC (Sect. 10.3). The
       calculated amount for the analyte in the QCS must be ± 20% of the expected
       value. If measured analyte concentrations are not of acceptable accuracy, check

                                  522-21

-------
             the entire analytical procedure to locate and correct the problem, or obtain another
             QCS.

   9.4   METHOD MODIFICATION QC REQUIREMENTS - The analyst is permitted to
         modify the GC injection technique, GC column and conditions, and MS conditions.

       9.4.1  Each time method modifications are made, the analyst must repeat the procedures
             of the IDC (Sect. 9.2) and verify that all QC criteria can be met in ongoing QC
             samples (Sect. 9.3).

       9.4.2  The analyst is also required to evaluate and document method performance for the
             proposed method modifications in real matrices that span the range of waters that
             the laboratory analyzes.  This additional step is required because modifications
             that perform acceptably in the IDC, which is conducted in reagent water, can fail
             ongoing method QC requirements in real matrices. This is particularly important
             for methods subject to matrix effects. If, for example, the laboratory analyzes
             finished waters from both surface and groundwater municipalities, this
             requirement can be accomplished by assessing precision and accuracy (Sects.
             9.2.2 and 9.2.3) in a surface water with moderate to high Total Organic Carbon
             (TOC) (e.g., 2 mg/L or greater) and a hard groundwater (e.g., 250 mg/L or greater
             as calcium carbonate).

       9.4.3  The results of Sections 9.4.1  and 9.4.2 must be appropriately  documented by the
             analyst and should be independently assessed by the laboratory's Quality
             Assurance (QA) officer prior to analyzing Field Samples.

          9.4.3.1   When implementing method modifications, it is the responsibility of the
                  laboratory to closely review the results of ongoing QC,  and in particular, the
                  results associated with the LFBs (Sect. 9.3.3), LFSMs (Sect. 9.3.7), FDs or
                  LFSMDs (Sect. 9.3.8),  CCCs (Sect. 9.3.2), and the IS area  counts (Sect.
                  9.3.5). If repeated failures are noted, the modification must be abandoned.

10. CALIBRATION AND STANDARDIZATION

   10.1  Demonstration and documentation of acceptable mass spectrometer (MS) tune and
         initial calibration is required before any samples are analyzed.  After the initial
         calibration is  successful, a CCC is required at the beginning and end of each period in
         which analyses are performed, and after every tenth Field Sample.  Verification of
         mass spectrometer tune must be repeated each time a major instrument modification is
         made or maintenance is performed, and prior to establishing or re-establishing an
         initial calibration.

   10.2  INITIAL CALIBRATION

       10.2.1 MS TUNE/MS TUNE CHECK - Operate the MS in the electron ionization (El)
             mode. Calibrate the mass and abundance scales of the MS with calibration

                                        522-22

-------
      compounds and procedures prescribed by the manufacturer with any
      modifications necessary to meet tuning requirements.  Inject 2 ng or less of the
      BFB solution (Sect. 7.2.4) into the GC/MS system.  Acquire a mass spectrum that
      includes data for m/z 45 to 180. The GC and MS parameters must be set such that
      a minimum of five scans (10 scans are recommended) are obtained during the
      elution of the BFB chromatographic peak  Use a single spectrum of the BFB
      peak, an average spectrum of the three highest points of the peak, or an average
      spectrum across the entire peak to evaluate the performance of the system.
      Appropriate background subtraction is permitted. If the BFB mass spectrum does
      not meet all  criteria in Table 1, the MS must be re-tuned and adjusted to meet all
      criteria before proceeding with the initial calibration.

10.2.2 ANALYTE  CALIBRATION - Operating conditions used during method
      development are described below. Data were obtained using SIM on a
      quadrupole MS system  using the GC column described in Section 6.10.1.
      Conditions different from those described may be used if QC criteria in Section 9
      are met. Different conditions include alternate GC columns, temperature
      programs, MS conditions, and injection techniques and volume, such as cold on-
      column and  direct injection port liners and/or large volume injection techniques.
      Equipment specifically  designed for alternate types of injections must be used if
      these options are selected. Full scan MS data may be used if the sensitivity
      achieved meets the data user's needs.

   10.2.2.1  GAS CHROMATOGRAPH (GC) CONDITIONS - Inject a l-|iL aliquot of
            each CAL prepared as in Section 7.2.2 into a hot, splitless injection port
            held at 200 °C with a split delay of 0.5 min. Temperature program the GC
            as follows: initial  oven temperature of 30  °C, hold for 1.0 min, ramp at
            7 °C/min to 90 °C, ramp at 20 °C/min to a final temperature of 200 °C and
            hold for 3.0 min, for a total run time of 18.07 min. Begin data acquisition at
            6.0 min.  During method development, the GC was operated in a constant
            flow rate mode at a rate of 1.0 mL per min. The MS ion source and the GC
            injector were both maintained  at 200 °C.

            NOTE: Using these conditions, all compounds of interest eluted before
            9 min.  However,  it is important to complete the  GC temperature program to
            ensure that co-extracted materials have eluted from the GC column prior to
            initiating the next analysis.

   10.2.2.2  MASS SPECTROMETER (MS) CONDITIONS - Any type of MS may be
            used as detailed in Sect. 6.10.4. Data shown in Section 17 (Tables 2-5) were
            obtained in the SEVI mode. Operation of the MS in this mode enhances
            sensitivity. However,  less mass spectral data are obtained for all peaks
            including the method analyte,  SUR, IS and any potential interferences.
            Because of this, and because the selected ions for the compounds of interest
            are very low masses that are likely to occur more frequently in interferences
            than most higher mass ions, the analyst should also rely on chromatography

                                 522-23

-------
               to reduce the possibility of false positives. It is highly recommended that a
               GC column at least 30 m in length with a film thickness of at least 1.4 um
               be used to provide adequate separation of the compounds of interest from
               possible interferences.  During method development, the MS was scanned in
               SIM mode for m/z 46, 78, and 80 (IS) at a rate of 0.36 s/scan in Segment 1
               (6 to 8 min). Segment 2 (8 to 18 min) was set to scan for m/z 58 and 88
               (TARGET), as well as 62, 64, and 96 (SUR) at 0.60 s/scan,  resulting in at
               least nine scans across each chromatographic peak. A minimum of five
               scans during the elution of each GC peak are required.  Ten scans across
               each GC peak are recommended. This requirement is applicable to both full
               scan and SIM analysis. Timing of the SIM segments must be set such that
               no chromatographic peak of interest begins or ends within 5 s of the
               beginning or end of the segment.  An example chromatogram obtained
               during method development using the instrumental conditions described in
               this section is shown in Figure 1.

   10.2.3  CALIBRATION CURVE - Use the GC/MS software to create a calibration curve
          for 1,4-dioxane using the IS technique. Concentrations may be calculated through
          the use of a linear or quadratic calibration curve. A weighted curve is permitted at
          the discretion of the analyst.  Because the SUR is added to all samples and
          standards at a single concentration, calibrate for the SUR using the average
          response factor.

   10.2.4  CALIBRATION ACCEPTANCE CRITERIA - When quantitated using the
          calibration curve, each calibration point, except the lowest point,  should calculate
          to be within 80-120% of its true value.  The lowest point should calculate to be
          within 60-140%  of its true value.  If these criteria cannot be met, the analyst will
          have difficulty meeting ongoing QC criteria. It is recommended that corrective
          action be taken to re-analyze the CALs, restrict the range of calibration, or select
          an alternate method of calibration.

10.3   CONTINUING CALIBRATION CHECK (CCC) - The CCC verifies the initial
      calibration at the beginning and end of each group of analyses, and after every tenth
      sample during analyses. In this context, a "sample" is considered to be a Field
      Sample.  The LRBs, LFBs, LFSMs, LFSMDs, FDs and CCCs are not  counted as
      samples.  The beginning CCC for each analysis batch must be at or below the MRL in
      order to verify instrument sensitivity prior to any analyses. Subsequent CCCs must
      alternate between medium and high concentration CALs.

   10.3.1  Inject an aliquot of the appropriate concentration CAL solution and analyze with
          the same conditions used during the initial calibration.

   10.3.2  Determine that the absolute area of the QI  of the IS has not changed by more than
          ± 50% from the average area measured during initial calibration, or more than
          ± 30% from the most recent CCC.  If the IS area has changed by more than this
                                    522-24

-------
             amount, remedial action must be taken (Sect. 10.3.4). Control charts are useful
             aids in documenting system sensitivity changes.

       10.3.3 Calculate the concentration of 1,4-dioxane and the SUR in the check standard.
             The calculated amount for 1,4-dioxane for medium and high level CCCs must be
             ± 30% of the true value. The calculated amount for the lowest calibration level
             for 1,4-dioxane, which must be at a concentration less than or equal to the MRL,
             must be within ± 50% of the true value.  If these criteria are not met, then all data
             for the problem analyte must be considered invalid, and remedial action
             (Sect. 10.3.4) should be taken.  This may require recalibration.  Any Field Sample
             extracts that have been analyzed since the last acceptable calibration verification
             should be re-analyzed after adequate calibration has been restored, with the
             following exception: if the CCC fails at the end of an analytical sequence because
             the calculated concentration of the target compound or SUR is greater than 130%
             (150% for the low-level CCC), and Field Sample extracts show no detection for
             the target compound, non-detects may be reported without re-analysis.

       10.3.4 REMEDIAL ACTION - Failure to meet CCC QC performance criteria may
             require remedial action. Major maintenance such as cleaning the ion source,
             cleaning the mass analyzer, replacing filament assemblies, replacing the GC
             column, etc., require returning to the initial calibration step (Sect.  10.2).

11. PROCEDURE

   11.1  This procedure may be performed manually or in an automated mode (Sect. 6.8.4)
         using a robotic or automatic sample preparation device. If an automatic system is used
         to prepare samples, follow the manufacturer's operating instructions, but all extraction
         and elution steps must be the same as in the manual procedure. Extraction and/or
         elution  steps may not be changed or omitted to accommodate the use of an
         automated system.

   11.2  Important aspects of this analytical procedure include proper preparation of laboratory
         glassware and sample containers (Sect. 4.1), as well as sample collection and storage
         (Sect. 8).  This section details the procedures for sample preparation, SPE using 2-g
         coconut charcoal or 0.4-g carbon cartridges as described in Sect. 6.8.1, and extract
         analysis.

         NOTE: The SPE cartridges and sodium sulfate drying materials described  in this
         section are designed as single use items and must be discarded after use.  They may
         not be refurbished for re-use in subsequent analyses.

         NOTE: Only purge and trap grade methanol may be used in this extraction (see
         Sect. 4.3).
                                        522-25

-------
11.3   SAMPLE PREPARATION

   11.3.1  All Field and QC samples (LRBs and LFBs) must contain the dechlorinating
          agent, sodium sulfite. Verify that Field Samples were checked for total chlorine
          upon sample receipt at the laboratory, and that no chlorine was present (Sect.
          8.3.1).  Add sodium sulfite to all LRBs and LFBs according to the instructions in
          Section 8, prior to fortification with the SUR and analyte.

   11.3.2  All Field and QC samples (LRBs and LFBs), must contain the anti-microbial
          agent sodium bisulfate. Verify that Field Samples were acidified to < pH 4 at the
          time of collection by checking with narrow range pH paper (this verification may
          take place at sample receipt if preferred by the laboratory). Add sodium bisulfate
          to all LRBs and LFBs according to the instructions in Section 8, prior to
          fortification with the SUR and analyte.

   11.3.3  If the sample bottles used contain a sample volume close to the  sample volume to
          be extracted (± 10%), mark the sample level on the outside of the sample
          container for subsequent measurement of the sample volume. After the sample
          has been extracted, determine the exact sample volume by filling the bottle to the
          mark with water and transferring to an appropriately sized graduated cylinder.
          Measure to within ± 5 mL for 100 mL samples, and to within ±  25 mL for
          500-mL samples. If a much larger sample has been collected than will be
          extracted, transfer either 100 mL or 500 mL of the sample to a clean glass
          container with a clean graduated cylinder. The sample volume will depend upon
          the SPE being used (Sects. 11.4, 11.5).

          Alternatively, the sample volume may be calculated by weighing the sample
          bottle before and after extraction (or sample transfer) and using  a density of
          l.Og/mL.

   11.3.4  ADDITION OF  SURROGATE (SUR) ANALYTE - Add an aliquot of the
          Surrogate Analyte Fortification Solution (Sect. 7.2.3.2) to all samples and mix by
          swirling the sample.  Addition of 2.5 uL of a 2000-ug/mL solution to a 500-mL
          sample will result in a concentration of 10 ug/L. Addition of 5  uL of a
          200-ug/mL solution to a 100-mL sample will also result in a concentration of
          10 ug/L.

   11.3.5  FORTIFICATION WITH METHOD ANALYTE - If the sample is an LFB,
          LFSM, or LFSMD, add the necessary amount of Analyte Fortification Solution
          (Sect. 7.2.3.1).  Swirl each sample to ensure all components are properly mixed.
          Addition of 2.5 uL of a 200-ug/mL solution to a 500-mL sample will result in a
          concentration of 1 ug/L. Addition of 5 uL of a 20-ug/mL solution to a 100-mL
          sample will also result in a concentration of 1 ug/L.

   11.3.6  Proceed with sample extraction Option 1  (Sect. 11.4) or Option  2 (Sect. 11.5).
                                     522-26

-------
11.4  SPE PROCEDURE OPTION 1; EXTRACTION OF 500-ML SAMPLES - Proper
      conditioning of the solid phase sorbent can have a marked effect on method precision
      and accuracy.  This section describes the SPE procedure using 2-g coconut charcoal
      cartridges and the transfer tube system and SPE manifold as described in Section 6.8.

    11.4.1  CARTRIDGE CONDITIONING

       11.4.1.1  Fill  the cartridge with approximately 3 mL of DCM, turn on the vacuum,
               and pull the solvent through, aspirating completely.

       11.4.1.2  Fill  the cartridge with approximately 3 mL of methanol, turn on the vacuum,
               and pull the solvent through, aspirating completely.

       11.4.1.3  Fill  the cartridge with approximately 3 mL of methanol and elute with
               vacuum to just above the top frit - not allowing the cartridge to go dry at the
               end. From this point forward, do not allow the cartridge to go dry.

       11.4.1.4  Fill  the cartridge with approximately 3 mL of reagent water, turn on the
               vacuum, and pull the water through, repeat five times, without allowing the
               cartridge to go dry in between washes or at the end.

    11.4.2  SAMPLE EXTRACTION - Attach a transfer tube from each sample bottle to
          each cartridge and then turn on the vacuum. Adjust the vacuum so  that the
          approximate flow rate is 10 mL/min.  After all the sample has passed through
          each SPE cartridge, detach the transfer tube and draw air through the cartridge for
          10 min at full vacuum. Turn off and release the vacuum. Proceed immediately
          with cartridge elution.

    11.4.3  CARTRIDGE ELUTION - Lift the extraction manifold top and insert a rack with
          collection tubes into the vacuum manifold tank to collect the extracts as they are
          eluted from the cartridges. Fill each cartridge with DCM. Pull enough of the
          solvent into the cartridge at low vacuum to soak the sorbent. Turn off the vacuum
          and vent the system. Allow the sorbent to soak in DCM for approximately 1 min.
          Apply a low vacuum and pull the DCM through the cartridge in a dropwise
          fashion into the collection tube.  Continue to add DCM to the cartridge as it is
          being drawn through until the volume of extract is about 9 mL, determined by the
          markings on the side of the collection tube.

    11.4.4  Remove collection tubes containing the extract from the vacuum manifold.
          Adjust the final volume as closely as possible to 10 mL with DCM. Small
          amounts of residual water from the sample container and the SPE cartridge may
          form a small immiscible layer with the extract. Read the volume on the collection
          tube at the level of the DCM layer. Add the IS.  Addition of 5 uL of a
          1000-ug/mL IS solution to a 10-mL extract will result in an IS concentration of
          500 ng/mL. Mix well. A vortex mixer is recommended. To eliminate residual
          water, add approximately 2 g  of anhydrous sodium sulfate.  Mix  well.  Transfer

                                     522-27

-------
          aliquots of the extract to autosampler vials, or other glass vials, for storage until
          analysis.

          NOTE: Experiments during method development indicated that this extract
          cannot be reliably concentrated by nitrogen evaporation because of the volatility
          of the method analyte.  Therefore, extract concentration to enhance sensitivity is
          not permitted.

11.5   SPE PROCEDURE OPTION 2; EXTRACTION OF 100-ML SAMPLES - Proper
      conditioning of the solid phase sorbent can have a marked effect on method precision
      and accuracy. This section describes the SPE procedure using Waters AC-2 Sep-Pak
      or Supelco Supelclean ENVI-Carb Plus cartridges with 60-mL reservoirs and the SPE
      manifold as described in Sect. 6.8.

   11.5.1  CARTRIDGE CONDITIONING

       11.5.1.1  Fill the cartridge with approximately 1 mL of DCM, turn on the vacuum,
               and pull the solvent through, aspirating completely.

       11.5.1.2 Place a reservoir over each cartridge and fill the cartridge with
               approximately 2 mL of methanol, turn on the vacuum, and pull the solvent
               through, aspirating completely.

       11.5.1.3  Fill the cartridge with approximately 2 mL of methanol and elute with
               vacuum to just above the top of the cartridge - not allowing the cartridge to
               go dry at the end. From this point forward, do not allow the cartridge to go
               dry.

       11.5.1.4 Fill the cartridge with approximately 3 mL of reagent water, turn on the
               vacuum, and pull the water through, without allowing the cartridge to go
               dry.

   11.5.2  SAMPLE EXTRACTION - Fill each reservoir with sample. Adjust the vacuum
          so that the approximate flow rate is 10 mL/min.  A 60-mL reservoir will require a
          second filling to complete sample loading. After all the sample has passed
          through each SPE cartridge, detach the reservoir and draw air through the
          cartridge for 10 min at full vacuum. Turn off and release the vacuum. Proceed
          immediately with cartridge elution.

   11.5.3  CARTRIDGE ELUTION - Lift the extraction manifold top and insert a rack with
          2-mL volumetric collection tubes into the extraction tank to collect the extracts as
          they are eluted from the cartridges. (It is not necessary to reverse the ENVI-Carb
          Plus cartridge as described in the manufacturer's literature.) Fill each cartridge
          with DCM. Pull enough of the solvent into the cartridge at low vacuum to soak
          the sorbent. Turn off the vacuum  and vent the system.  Allow the sorbent to soak
          in DCM for approximately 1  min. Apply a low vacuum and pull the DCM

                                     522-28

-------
             through the cartridge in a dropwise fashion into the collection tube. Continue to
             add DCM to the cartridge as it is being drawn through until the volume of extract
             is about 1.5 mL, estimated by the volumetric mark on the side of the collection
             tube.

       11.5.4 Remove collection tubes containing the extracts from the vacuum manifold.
             Adjust the final volume as closely as possible to 2 mL with DCM.  Small amounts
             of residual water from the sample container and the SPE cartridge may form a
             small immiscible layer with the extract. Read the volume on the collection tube at
             the level of the DCM layer. Add the IS. Addition of 10 uL of a 100-ug/mL IS
             solution to a 2-mL extract will result in an IS concentration of 500 ng/mL. Mix
             well.  A vortex mixer is recommended. To eliminate residual water, pass the
             2-mL extracts through a small column containing a minimum of 0.4 g of dry
             anhydrous sodium sulfate.  A fritted 3-mL polypropylene column or a 5-8 mm
             diameter disposable pipette with a small piece of glass wool may be used. Collect
             the extracts from the drying column directly in 2-mL autosampler vials for storage
             until analysis.

             NOTE: Experiments during method development indicated that this extract
             cannot be reliably concentrated by nitrogen evaporation because of the volatility
             of the method analyte. Therefore, extract concentration to enhance sensitivity is
             not permitted.

   11.6  Analyze an aliquot of the sample extract prepared using Option 1 or 2, with the
         GC/MS system under the same conditions used for the initial and continuing
         calibrations (Sect. 10).

12.  DATA ANALYSIS AND CALCULATIONS

   12.1  IDENTIFICATION OF ANALYTES - At the conclusion of data acquisition, use the
         same software that was used in the calibration procedure to identify peaks in
         predetermined retention time windows of interest. Use the data system software to
         examine the ion abundances of components of the chromatogram.  Identify a sample
         component by comparison of its SIM (or full scan) mass spectrum to a SIM (or full
         scan) reference spectrum in the user-created database.  The GC retention time of a
         method analyte should be within one to two sec of the retention time observed for that
         same compound in the most recently analyzed CCC standard.  Ideally, the width of the
         retention time window should be based upon measurements of actual retention time
         variations of standards over the course of a day.  Three times the standard deviation of
         a retention time can be used to calculate a suggested window size for an analyte.
         However, the experience of the analyst should weigh heavily in the interpretation of
         the chromatogram.  When this method is performed in the SEVI mode, verification of
         retention times is particularly important because less mass spectral information is
         being collected.
                                        522-29

-------
   12.2  IDENTIFICATION VERIFICATION USING ION RATIOS - When the QI of
         1,4-dioxane (m/z 88) is observed at the correct retention time, verify that it is truly
         present by verifying that the ratio of the confirmation ion (m/z 58) to the QI  (m/z 88) is
         within an absolute ±20% of the ratio observed in a mid-range calibration standard. For
         example, if the ratio in the standard  is 70%, then the ratio in a Field Sample  must be
         within 50-90%  Ratios must be determined using the integrated areas for the QI
         and confirmation ion from their respective mass chromatograms. Using relative
         abundances obtained directly from the mass spectrum, whether or not it has  undergone
         background subtraction,  is not acceptable.  Variations in system software and analyst
         subjectivity in selecting background spectra for subtraction can affect the validity of
         the ratio calculation when the relative abundance from the spectrum is used. Using a
         raw spectrum (not corrected for background) can also affect the validity of the ratio
         calculation.

         Use similar procedures for identification of the IS and SUR. Suggested QIs and
         confirmation ions for these compounds may be found in Table 2 (Sect. 17).

   12.3  Calculate analyte and SUR concentrations using the multi-point calibration established
         in Section 10.2. Do not use daily CCC data to quantitate 1,4-dioxane or the SUR in
         samples.  The integrated abundances of the QIs of the analyte, SUR and IS should be
         used for all calculations. Adjust the final analyte concentrations to reflect the actual
         sample volume as determined in Sect. 11.3.3.  Field Sample extracts that require
         dilution should be treated as described in Section 12.4.

   12.4  EXCEEDING CALIBRATION RANGE - An analyst must not extrapolate  beyond the
         established calibration range. If an analyte result exceeds the range of the initial
         calibration curve, the extract may be diluted with DCM, with the appropriate amount
         of IS added to match the original concentration,  and the diluted extract injected.
         Acceptable SUR performance (Sect. 9.3.6.1) should be determined from the undiluted
         sample extract.  Incorporate the dilution factor into final concentration calculations.
         The resulting sample should be documented as a dilution, and the MRL adjusted
         accordingly.

   12.5  Calculations must utilize all available digits of precision, but final reported
         concentrations should be rounded to an appropriate number of significant figures (one
         digit of uncertainty), typically two, and not more than three significant figures.

         NOTE: Some data in Section 17 of this method are reported with more than two
         significant figures.  This is done to better illustrate the method performance  data.

13. METHOD PERFORMANCE

   13.1  Method performance data presented in Section 17 were obtained with a Thermo
         Finnigan Trace DSQ GC/MS system.  Any GC/MS system that meets all of the QC
         requirements of the method may be  used.
                                        522-30

-------
   13.2  PRECISION, ACCURACY, AND MINIMUM REPORTING LEVELS - Single
         laboratory DLs and LCMRLs are presented in Table 3. Single laboratory accuracy
         and precision data from both fortified reagent water and fortified drinking water
         matrices are presented in Tables 4 and 5, respectively.  Two of the matrices tested
         were selected specifically as potentially difficult matrices because of either their
         relatively high TOC or high mineral content (hardness). There was no apparent
         difference in method performance when results from reagent water studies were
         compared to these drinking water matrices.

   13.3  METHOD PERFORMANCE IN SAMPLES WITH HIGH CONCENTRATIONS OF
         1,1,1-TRICHLORETHANE - Because of its widespread use as a chlorinated solvent
         stabilizer, 1,4-dioxane may be found in the presence of high concentrations of
         1,1,1-trichloroethane. The method was tested in a high TOC water matrix with up to
         500 ug/L of 1,1,1-trichloroethane added as a co-contaminant.  No adverse affect was
         observed, and the method performed well within QC limits (data not shown).

   13.4  SAMPLE STORAGE STABILITY STUDIES

       13.4.1 AQUEOUS SAMPLES - Chlorinated drinking water samples from a surface
             water source were fortified with 1,4-dioxane and preserved  and stored as required
             in Section 8. Samples were extracted and analyzed in replicate (n=7) on day 0,
             and at five additional times points up to and beyond 28 days. Data from these
             analyses validate the 28 day holding time, and are presented in Figure 2.

       13.4.2 EXTRACTS - Sample extracts stored at -5 °C and protected from light were
             analyzed in replicate (n=7) on day 0, and at three  additional time points up to and
             beyond 28 days. Data from these analyses validate the 28 day holding time, and
             are presented in Figure 3.

   13.5  MULTIPLE LABORATORY DEMONSTRATION - The performance of this method
         was demonstrated in two additional laboratories, with  results similar to those reported
         in Section 17. The authors wish to acknowledge the assistance of the analysts and
         laboratories listed below for their participation in the multi-lab demonstration.

       13.5.1 Mr. Alan Zaffiro of Shaw Environmental and Infrastructure, Inc. under contract to
             the USEPA Office of  Ground Water and Drinking Water Technical Support
             Center, Cincinnati, OH.

       13.5.2 Dr. Peggy Knight and Ms. Megan Pickett of the USEPA Region 10 Laboratory,
             Port Orchard, WA.

14. POLLUTION PREVENTION

   14.1  This method utilizes SPE to extract the analyte from water.  It requires the use of very
         small volumes of organic solvent and very small quantities of pure analyte, thereby
         minimizing the potential hazards to both the analyst and the environment as compared

                                       522-31

-------
         to the use of large volumes of organic solvents in conventional liquid-liquid
         extractions.

    14.2  For information about pollution prevention that may be applicable to laboratory
         operations, consult "Less is Better: Laboratory Chemical Management for Waste
         Reduction" available from the American Chemical Society's Department of
         Government Relations and Science Policy, 1155 16th Street N.W., Washington, D.C.,
         20036 or on-line at http://membership.acs.org/c/ccs/pub_9.htm. (accessed September
         2008).

15. WASTE MANAGEMENT

    15.1  The analytical procedures described in this method generate relatively small amounts
         of waste since only small amounts of reagents and solvents are used.  The matrices of
         concern are finished drinking water or source water.  Waste management practices
         must be conducted consistent with all applicable rules and regulations, and
         laboratories must protect the air, water, and land by minimizing and controlling all
         releases from fume hoods and bench operations. Compliance is required with any
         sewage discharge permits and regulations, particularly the hazardous waste
         identification rules and land disposal restrictions.

16. REFERENCES

1.   Winslow, S.D., Pepich, B.V., Martin, J.J., Hallberg, G.R., Munch, D.J., Frebis, C.P.,
    Hedrick, EJ. and R.A. Krop, Statistical Procedures for Determination and Verification of
   Minimum Reporting Levels for Drinking Water Methods. Environ. Sci. Technol., 40, (2006)
    281-288.

2.   Glaser, J.A., Foerst, D.L., McKee, G.D., Quave, S.A. and W.L. Budde, Trace Analyses for
    Wastewaters. Environ. Sci. Technol.. 15, (1981) 1426-1435.

3.   Kadokami, K., Koga, M. and Otsuki, Gas Chromatography/Mass Spectrometric
   Determination of Traces ofHydrophilic and Volatile Organic Compounds in Water after
   Preconcentration with Activated Carbon. Analytical Sciences, 6, (1990) 843-849.

4.   Isaacson, C., Mohr, T.K.G. and J.A. Field, Quantitative Determination of 1,4-Dioxane and
    Tetrahydrofuran in Groundwater by Solid Phase Extraction GC/MS/MS. Environ. Sci.
    Technol.. 40, (2006) 7305-7311.

5.   USEPA, 2006 Edition of the Drinking Water Standards and Health Advisories. APA #822-R-
    06-013.

6.   Carcinogens - Working With Carcinogens. Department of Health, Education, and Welfare,
    Public Health Service, Center for Disease Control, National Institute for Occupational Safety
    and Health, Publication No. 77-206, Aug. 1977.
                                        522-32

-------
7.  OSHA Safety and Health Standards, General Industry. (29CFR1910), Occupational Safety
   and Health Administration, OSHA 2206, (Revised, July 1, 2001).

8.  Safety in Academic Chemistry Laboratories. American Chemical Society Publication,
   Committee on Chemical  Safety, 7th Edition. Information on obtaining a copy is available at
   http://membership.acs.org/C/CCS/pub_3.htm. (accessed September 2008).  Also available by
   request at OSS@acs.org
                                        522-33

-------
17.TABLES, DIAGRAMS, FLOWCHARTS AND DEMONSTRATION DATA
Table 1.   Bromofluorobenzene (BFB) Tune Verification Criteria
Mass (m/z)
50
75
95
96
173
174
175
176
177
Relative Abundance Criteria
15-40% of mass 95
3 0-80% of mass 95
Base peak, 100% relative abundance
5-9% of mass 95
<2%ofmass 174
>50%ofmass95
5-9% of mass 174
>95% but <101% of mass 174
5-9% of mass 176
Table 2.   Retention Times and Quantitation Ions (QIs)
Compound
1,4-dioxane
l,4-dioxane-J8(SUR)
THF-J8 (IS)
Retention Time (min)
8.85
8.77
6.68
SIM Ions (m/z)
58a, 88
62, 64, 96
46, 78, 80
 Note: Suggested quantitation ion in bold.
 a. Ion traps may give a confirmation ion of m/z 57 for 1,4-dioxane instead of the typical m/z 58
   in quadrupole spectra and in many mass spectral libraries. If the instrument has been
   calibrated and passes bromofluorobenzene (BFB) criteria, then m/z 57 may be used as the
   confirmation ion on ion trap instruments.
 Table 3.  Lowest Concentration Minimum Reporting Level (LCMRL) and Detection
          Limit (DL) Calculated for Each Extraction Option
Compound/Extraction Option
1,4-dioxane (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC-2 Sep-Pak cartridge)
LCMRL
(HS/L)
0.047
0.036
DL (ug/L)
0.026a
0.0206
a. Calculated from replicates fortified at 0.040 ug/L, n=7.
b. Calculated from replicates fortified at 0.030 ug/L, n=8.
                                       522-34

-------
Table 4.   Demonstration of Method Performance in Reagent Water Fortified with
          1,4-Dioxane at Three Concentrations
Compound/Extraction Option
1,4-dioxane (500 ml w/2-g Coconut charcoal)
l,4-dioxane-J8 (SUR) (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC-2 Sep-Pak cartridge)
l,4-dioxane-J8(SUR) (100 ml w/Waters AC-2 Sep-Pak
cartridge)

1,4-dioxane (500 ml w/2-g Coconut charcoal)
l,4-dioxane-J8 (SUR) (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC-2 Sep-Pak cartridge)
l,4-dioxane-6?8 (SUR) (100 ml w/Waters AC-2 Sep-Pak
cartridge)

1,4-dioxane (500 ml w/2-g Coconut charcoal)
l,4-dioxane-J8 (SUR) (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC-2 Sep-Pak cartridge)
l,4-dioxane-6?8 (SUR) (100 ml w/Waters AC-2 Sep-Pak
cartridge)
Reagent Water Fortified at
0.030 or 0.040 ug/L
(n=7 or 8) a
Mean %
recovery
110
92
110
102
RSD (%)
19
3.5
20
3.8
Reagent Water Fortified at
1.0 ug/L (n=6)
Mean %
recovery
98.0
97.1
101
105
RSD (%)
6.2
5.2
4.2
4.8
Reagent Water Fortified at
10.0 ng/L (n=7)
Mean %
recovery
93.9
90.1
95.2
99.3
RSD (%)
2.8
2.8
2.9
3.0
a. Samples (500 mL) extracted on 2-g Coconut charcoal cartridges were fortified at 0.040 ug/L,
  n=7. Samples (100 mL) extracted on Waters AC-2 Sep-Pak cartridges were fortified at 0.030
  ug/L, n=8.
                                       522-35

-------
Table 5.  Demonstration of Method Performance in Drinking Water Matrices Using Each
          of the Solid Phase Extraction (SPE) Options; Matrix Samples Fortified at 1.0
          ug/L (n=7 for each matrix)
Compound/Extraction Option
1,4-dioxane (500 ml w/2-g Coconut charcoal)
l,4-dioxane-J8 (SUR) (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC -2 Sep-Pak cartridge)
1,4-dioxane-dg (SUR) (100 ml w/Waters AC -2 Sep-Pak
cartridge)

1,4-dioxane (500 ml w/2-g Coconut charcoal)
l,4-dioxane-J8 (SUR) (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC -2 Sep-Pak cartridge)
1,4-dioxane-dg (SUR) (100 ml w/Waters AC -2 Sep-Pak
cartridge)

1,4-dioxane (500 ml w/2-g Coconut charcoal)
l,4-dioxane-J8 (SUR) (500 ml w/2-g Coconut charcoal)
1,4-dioxane (100 ml w/Waters AC -2 Sep-Pak cartridge)
1,4-dioxane-dg (SUR) (100 ml w/Waters AC -2 Sep-Pak
cartridge)
Surface Water
Mean %
recovery
99.0a
100
97.0b
98.5
RSD (%)
4.9
2.5
4.6
2.5
Surface Water (high in
TOC)C
Mean %
recovery
102
99.8
98.5
101
RSD (%)
3.5
2.4
5.6
4.2
Groundwater (high in
mineral content) d
Mean %
recovery
95. 9e
98.2
1011
104
RSD (%)
2.1
2.5
3.3
5.9
a. Percent recovery after correction for matrix background of 0.66 ug/L.
b. Percent recovery after correction for matrix background of 0.42 ug/L.
  Note: Groundwater used for these Laboratory Fortified Blanks (LFBs) was collected from the
  same source as the 500 mL LFBs using coconut charcoal, but was collected on different days,
  thus the difference in matrix background concentration.
c. TOC measured at 4.950 mg/L.
d. Hardness measured at 289 mg/L as calcium carbonate.
e. Percent recovery after correction for matrix background of 0.070 ug/L.
f. Percent recovery after correction for matrix background of 0.080 ug/L.
                                        522-36

-------
Table 6.   Initial Demonstration of Capability (IDC) and Quality Control (QC) Requirements (Summary)
Method
Reference
Sect. 9.2.1
Sect. 9.2.2
Sect. 9.2.3
Sect. 9.2.4
Sect. 9.2.5
& 9.3.9
Requirement
Initial Demonstration of
Low Background
Initial Demonstration of
Precision (IDP)
Initial Demonstration of
Accuracy (IDA)
Minimum Reporting
Limit (MRL)
Confirmation
Calibration Confirmation,
Quality Control Sample
(QCS)
Specification and Frequency
Analyze LRB prior to any other IDC steps. When a new
lot of SPE media is obtained, verify that background is
at acceptable limits.
Analyze 4-7 replicate LFBs fortified near the midrange
calibration concentration.
Calculate average recovery for replicates used in IDP.
Fortify, extract and analyze 7 replicates at the proposed
MRL concentration. Calculate the mean, standard
deviation and HRpiR. Confirm that the upper and lower
limits for the Prediction Interval of Result (Upper PIR,
and Lower PIR, Sect. 9.2.4.2) meet the recovery criteria.
Analyze a standard from a second source (QCS) to
verify the initial calibration curve.
Acceptance Criteria
Demonstrate that the method analyte is < 1/3 the
MRL, and that possible interferences from
extraction media do not prevent the identification
and/or quantification of the method analyte, SUR
or IS.
Note: This includes the absence of interferences at
both the QIs and confirmation ions at the RTs of
interest.
%RSDmustbe<20%
Mean recovery + 20% of true value
Upper PIR < 150%
Lower PIR > 50%
± 20% of the expected value.
 NOTE: Table 6 is intended as an abbreviated summary of QC requirements provided as a convenience to the method user. Because the information has been
 abbreviated to fit the table format, there may be issues that need additional clarification, or areas where important additional information from the method text
 is needed.  In all cases, the full text of the QC in Section 9 supersedes any missing or conflicting information in this table.
                                                                522-37

-------
Table 7.   Ongoing Quality Control (QC) Requirements (Summary)
  Method
  Reference
Requirement
Specification and Frequency
Acceptance Criteria
  Sect. 8.3
Sample Holding Time
28 days with appropriate preservation and storage as
described in Sections 8.1-8.3.
Sample results are valid only if samples are
extracted within sample hold time.
  Sect. 8.4
Extract Holding Time
28 days stored at -5 °C and protected from light
Sample results are valid only if extracts are
analyzed within extract hold time.
  Sect. 9.3.1
Laboratory Reagent Blank
(LRB)
One LRB with each extraction batch of up to 20 Field
Samples.
Demonstrate that the method analyte concentration
is < V3 the MRL, and confirm that possible
interferences do not prevent quantification.  If the
background concentration exceeds V3 the MRL,
results for the extraction batch are invalid.
  Sect. 9.3.3
Laboratory Fortified
Blank (LFB)
One LFB is required for each extraction batch of up to
20 Field Samples. Rotate the fortified concentrations
between low, medium, and high amounts.
Results of LFB analyses at medium and high
fortifications must be 70-130% of the true value for
the analyte and SUR. Results of the low-level LFB
must be 50-150% of the true value.
  Sect. 9.3.5
Internal Standard (IS)
Compare IS area to the average IS area in the initial
calibration and the most recent CCC.
Peak area counts for all injections must be within
+ 50% of the average peak area calculated during
the initial calibration and + 30% from the most
recent CCC. If the IS does not meet this criterion,
target analyte results are invalid.  Consult Sect.
9.3.5 for further information.
  Sect. 9.3.6
Surrogate(SUR)
Standards
The SUR standard added to all calibration standards and
samples, including QC samples. Calculate SUR
recoveries.
SUR recovery must be 70-130% of the true value.
If a SUR fails this criterion, report all results for
sample as suspect/SUR recovery.
  Sect. 9.3.7
Laboratory Fortified
Sample Matrix (LFSM)
Analyze one LFSM per extraction batch (of up to 20
Field Samples) fortified with the method analyte at a
concentration close to but greater than the native
concentration.  Calculate  LFSM recoveries.
See Sect. 9.3.7.3 for instructions on the
interpretation of LFSM results.
                                                                   522-38

-------
Table 7. Ongoing Quality Control (QC) Requirements (Summary) (Continued)
  Sect. 9.3.8
Laboratory Fortified
Sample Matrix Duplicate
(LFSMD) or Field
Duplicates (FD)
Extract and analyze at least one FD or LFSMD with
each extraction batch of up to 20 Field Samples. A
LFSMD may be substituted for a FD when the
frequency of detects for 1,4-dioxane are low.  Calculate
RPDs.
Method analyte RPDs for the LFSMD or FD should
be < 30% at mid and high levels of fortification and
< 50% near the MRL. Failure to meet this criterion
may indicate a matrix effect.
  Sect. 9.3.9
Quality Control Sample
(QCS)
Analyze QCS during the IDC, each time CAL solutions
are prepared.  A QCS must be analyzed at least
quarterly.	
Results must be 80-120% of the expected value.
  Sect. 10.2
Initial Calibration
Use IS calibration technique to generate a linear or
quadratic calibration curve. The number of standards
required is determined by the calibration range (Sect.
7.2.2). Check the calibration curve as described in
Section 10.2.4.
When each calibration standard is calculated as an
unknown using the calibration curve, the result
should be 80-120% of the true value for all except
the lowest standard, which should be 60-140% of
the true value. If this criterion is not met, reanalyze
CALs, select a different method of calibration or
recalibrate over a shorter range.
  Sects. 10.1
  and 10.2.1
MS Tune Check
Analyze BFB to verify MS tune each time the
instrument is calibrated.
Criteria are given in Table 1.
  Sect. 10.3
Continuing Calibration
Check (CCC)
Verify initial calibration by analyzing a calibration
standard at the beginning of each analysis batch prior to
analyzing samples, after every 10 samples, and after the
last sample. The first CCC daily must be at or below the
MRL. Subsequent CCCs alternate between medium and
high concentrations.

Low CCC - at or below the MRL concentration
Mid CCC - near midpoint in initial calibration curve
High CCC - near the highest calibration standard.
                                                                                             Low: + 50% of true value
                                                                                             Mid: +30% of true value
                                                                                             High: +30% of true value
  Note: Table 7 is intended as an abbreviated summary of QC requirements provided as a convenience to the method user. Because the information has been
  abbreviated to fit the table format, there may be issues that need additional clarification, or areas where important additional information from the method text
  is needed. In all cases, the full text of Sections 8-10 in the method supersedes any missing or conflicting information in this table.
                                                                   522-39

-------
RT: 5.84-9.32
    100000-
     50000-
  —
  c
         0-
            m/z88
                                                                    1,4-dioxane
                                                                      8.864
    150000-
    100000r
     50000-
  I
m/z96
                                                1,4-diox-d8
                                                 8.786
200000^
150000!
100000!
50000|
6.
m/z 46 THF-d8 ^J
6.722

Jy
845*



                      6.722
300000-E
200000!
100000!
ru
UJMR* 8-786
ft A ^ 8.864
I II /111
J \J L ~A~ _ „ rJ " V „
6.0         6.5         7.0        7.5        8.0
                               Time(min)
                                                                   8.5
                                                                  9.0
Figure 1. Reconstructed total ion current chromatogram and mass chromatograms for THF-J8 (IS),
l,4-dioxane-J8 (SUR), and 1,4-dioxane at 0.5 ug/mL each (the standard is equivalent to an extract of
a 10 jig/L aqueous sample).  * Peak at 6.845 min is chloroform, a chemical present in DCM.
                                           522-40

-------
                              Stability of 1,4-Dioxane in Drinking Water
       120.0%


       100.0%


        80.0%
    §   60.0% H
    a:
        40.0%
        20.0%
         0.0%
99.2%
              95.3%
103.2%

  f
99.4%
n
98.9%
                                                  14

                                                 days
                                             21
                               28
Figure 2.  Stability of 1,4-dioxane in preserved drinking water stored at 4°C in the dark, over a 35-
day time period. Replicate samples (n = 7) were fortified to a concentration of 1 ug/L 1,4-dioxane.
Matrix blank data were used to correct for native analyte concentrations.
120.0% -,
100.0% -
. 80.0% -
2"
0
8 60.0% -
cc
S.O
° 40.0% -
20.0%
0.0% -
Stability of 1 ,4-Dioxane in Extracts




93.9%








92.0%








91.2%








0 14 28




days
Figure 3.  Stability of 1,4-dioxane in sample extracts stored at -5°C, in the dark, over a 42-day time
period. Replicate samples (n = 7) were fortified to a concentration of 10 jig/L 1,4-dioxane.
                                           522-41

-------