THE ENVIRONMENTAL TECHNOLOGY VERIFICATION
                                              PROGRAM
              EPA
           VS. H. vir».«,»ul Fr>tcdi»
                                         Baireiie
                                     The Business of Innovation
      TECHNOLOGY TYPE:   PARTICULATE METALS MONITOR
      APPLICATION:
MONITORING OF METALS IN AEROSOL PARTICLES
IN AMBIENT AIR
      TECHNOLOGY NAME:  Xact 625
COMPANY:

ADDRESS:
      WEB SITE:
      E-MAIL:
                                 Cooper Environmental Services
10180 SW Nimbus Ave. Suite J6   PHONE: (503) 670-8127
Portland, OR 97223

http://cooperenvironmental.com/
info@cooperenvironmental.com
                           ETV Joint Verification  Statement

   The U.S. Environmental Protection Agency (EPA) has established the Environmental Technology Verification
   (ETV) Program to facilitate the deployment of innovative or improved environmental technologies through
   performance verification and dissemination of information. The goal of the ETV Program is to further
   environmental protection by accelerating the acceptance and use of improved and cost-effective technologies.
   ETV seeks to achieve this goal by providing high-quality, peer-reviewed data on technology performance to
   those involved in the design, distribution, financing, permitting, purchase, and use of environmental
   technologies. Information and ETV documents are available at www.epa.gov/etv.

   ETV works in partnership with recognized standards and testing organizations, with stakeholder groups
   (consisting of buyers, vendor organizations, and permitters), and with individual technology developers.  The
   program evaluates the performance of innovative technologies by developing test plans that are responsive to
   the needs of stakeholders, conducting field and laboratory tests (as appropriate), collecting and analyzing data,
   and preparing peer-reviewed reports. All evaluations are conducted in accordance with rigorous quality
   assurance (QA) protocols to ensure that data of known and adequate quality are generated and that the results
   are defensible.
t The license for the Xact 625 technology was given by Pall Corporation to Cooper Environmental Services LLC after completion of this
verification. The box above was revised on April 15, 2013 to reflect this change. The product name remains the same. The remainder of this
verification statement refers to the company name which was applicable when the technology was originally verified.

-------
The Advanced Monitoring Systems (AMS) Center, one of six verification centers under ETV, is operated by
Battelle in cooperation with EPA's National Risk Management Research Laboratory.  The AMS Center
evaluated the performance of a field deployable X-ray fluorescence (XRF) analyzer for determining metals in
particulate matter in ambient air.  This verification statement provides a summary of the test results for Pall
Corporation's1 Xact 625 Particulate Metals Monitor.

VERIFICATION TEST DESCRIPTION

This verification evaluated the ability of the Xact 625 to provide reliable quantitative measurements of metals in
particulate matter having aerodynamic diameter less than 10 micrometers (i.e., PMio), during continuous
unattended operation. The verification was conducted over several weeks in the winter of 2011-2012 and
involved continuous operation of an Xact 625 unit at an existing ambient air monitoring site in Marietta,  Ohio.
The verification was supported by EPA's Region 5 and Office of Air Quality Planning and Standards and was
carried out with the help of the Ohio Environmental Protection Agency. A total of 23 metals were reported
hourly by the Xact 625 over a field period of approximately 80 days.  Daily averages of Xact 625 metals readings
that exceeded the daily quantitation limits (QLs) of the analyzer were compared to corresponding results
determined by EPA reference sampling and analysis by inductively coupled plasma mass spectrometry (ICP/MS).
Quantitative comparisons of Xact 625 results to reference results could be made for six metals  [calcium (Ca),
copper (Cu), manganese (Mn), lead (Pb), selenium (Se), and zinc (Zn)], using reference measurements obtained
over approximately 65 days within the 80-day period of continuous Xact 625 operation.

The following are the parameters on which Xact 625 performance was evaluated:

    Comparability - The comparability of Xact 625 results to reference results was determined for each metal by
    examination of the slope and intercept of a linear regression of the data using the ICP/MS results as the
    independent variable and the Xact 625 daily averages as the dependent variable.

    Correlation - The degree of correlation between the Xact 625 results and the reference method results was
    determined for each  metal by the coefficient of determination (r2) of the linear regression between the
    reference method results and the corresponding Xact 625 daily averages.

   Bias - The bias of the Xact 625 readings was calculated as the percent difference  of each daily average Xact
    625 reading relative  to the corresponding daily average reference method result.

   Data Completeness - The completeness of Xact 625 data was calculated as the percentage of the field period
    for which valid hourly data were reported,  and as the percentage  of all daily reference measurement periods
    during which the Xact 625 reported at least 12 hours of monitoring data (i.e., produced data for at least half
    of the reference monitoring period).

    Operational Factors - Performance factors such as maintenance requirements,  ease of use, and effectiveness
    of data acquisition were assessed from observations by EPA and  Ohio EPA staff.

Battelle and EPA QA staff conducted separate  technical systems audits (TSAs) of the field operations, and
Battelle QA staff conducted a TSA of the reference analysis laboratory and a data quality audit of the test data.
This verification statement, the full report on which it is based, and the test/QA plan for this verification are
available at www.epa.gov/etv/centers/centerl.html.

TECHNOLOGY DESCRIPTION

The following description of the Xact 625 is based on information provided by the vendor and was not verified in
this evaluation: The Xact 625 determines metals in airborne PM10 by XRF, in which X-rays from a source cause
ejection of an electron from an inner electronic shell of a metal atom  in a sample. The vacancy in the inner shell
is filled by an electron from an outer shell, with resulting emission of an X-ray of wavelength longer than that of
the original excitation and characteristic of the metal in question. The Xact 625  uses energy-dispersive XRF
(EDXRF) and a low power X-ray source.  EDXRF provides rapid acquisition of the entire X-ray spectrum so that

-------
many elements can be detected within a few seconds. To monitor metals in atmospheric PM10, the Xact 625 uses
an automated moveable filter tape, in which sample air is drawn through a small spot on the tape, collecting and
concentrating PMio onto that spot. The tape then advances, placing the collected sample spot in the X-ray
excitation and analysis section of the monitor and initiating particulate sampling onto a previously unexposed spot
on the tape.  The sequence of sampling and analysis can continue automatically, limited only by the supply of
filter tape. The duration of sample collection at each spot can be set at a constant interval, or varied to maintain
detection performance when atmospheric PM10 levels vary. The Xact 625 samples ambient air at a constant flow
rate of 16.7 L/min (i.e.,  1 m3/hr).  The Xact 625's sample inlet is designed to provide uniform sample deposition,
and the instrument analyzes approximately 90% of the sample spot area to minimize effects of sample
inhomogeneity.  The Xact 625 incorporates sensors for temperature and atmospheric pressure, and uses those data
to maintain a constant volumetric sample flow and consequently an accurate PM10 inlet size cut.  For this
verification the vendor of the Xact 625 programmed the monitor to perform the following automatic internal
quality control (QC) checks on a daily basis to assure data quality:
•   Internal energy alignment check, by XRF analysis of a Cu rod, conducted over a 15-minute period starting at
    midnight each day,
•   Upscale rod check, by XRF analysis of a metal rod containing chromium (Cr),  Pb, and cadmium (Cd),
    conducted once per day over a 15-minute period,
•   Flow check, conducted at the same time as the upscale rod check, to determine the Xact 625 sample air flow
    by insertion of a second mass flow meter into the flow path,
•   Palladium rod stability check by XRF analysis of a palladium rod in every ambient sample analysis.

VERIFICATION RESULTS

Of the 23 ambient PMio metals determined by the Xact 625 in this verification, 19 were also determined by the
ICP/MS analysis of reference samples. Of those 19 metals, Xact 625 results for 12 metals were almost always
below the respective QLs of the Xact 625, and consequently no comparisons could be made to the reference
results. Also the Xact 625 reported a nearly constant and unrealistic ambient concentration of molybdenum due
to an incorrect energy window for that metal in the XRF analysis. The vendor of the Xact 625  indicates that this
problem can be avoided in the future by narrowing the energy window. Table  VS-1 presents a summary of the
quantitative performance results of the Xact 625 for the remaining six metals, showing the  performance
parameters and resulting performance metrics.

Table VS-1 indicates that the daily average Xact 625 results were highly correlated and in close quantitative
agreement with ICP/MS analysis results for most of the six metals, and that the Xact 625 achieved data
completeness of over 95%. The regression results for Cu reflect the fact that the ambient Cu concentrations were
often near the detection limit of the ICP/MS analysis and the QL of the Xact 625.  In addition, investigations
conducted after the verification test suggest that the regression results for Cu and the relatively high bias results
for Pb may both have been affected by the background levels  of these metals in the Xact 625's blank filter tape.

The Xact 625 required minimal operator attention during testing, with routine maintenance consisting of changing
the sampling tape approximately every 2 weeks. The analyzer conducted several automated internal QC checks
on a daily basis, and provided readily understandable data files that distinguished QC from ambient monitoring
results and included analyzer flow and internal readings.  The internal QC check results of the Xact 625 showed
that the relative standard deviation of the  palladium internal standard readings  determined in every Xact 625
analysis was 2.0%, and the instrument's average flow error was -0.69% (± 0.004%) where the number in
parentheses is the standard deviation of the daily flow check results. The average accuracy relative  to upscale rod
check standards analyzed daily was 100.2% (±  0.004%) for Cr, 100.6% (± 0.013%) for Cd, and 99.8% (±
0.007%) for Pb, where accuracy of 100% indicates perfect agreement with the  upscale rod check standard.
However, the test results for Cu and Pb strongly indicate that analysis of blank portions of the  filter tape should be
a component of the Xact 625's automated internal  QC checks.

-------

Table VS-1. Quantitative Performance Results for the Xact 625
Performance
Parameter
Comparability
Correlation
Bias
Data Completeness
Metal

Ca
Cu
Mn
Pb
Se
Zn

Ca
Cu
Mn
Pb
Se
Zn

Ca
Cu
Mn
Pb
Se
Zn

All
Metric
Slope
(± 95% CI)
0.82 (± 0.04)
0.30 (±0.1 7)
1.0 (± 0.01)
1.1 (±0.07)
0.99 (±0.11)
0.97 (± 0.03)
Intercept
(± 95% CI)
(ng/m3)
-30.6 (± 11.9)
2.1 (±0.65)
-0.09 (±2. 13)
1.6 (± 0.30)
0.01 (± 0.22)
-3.0 (± 0.98)
Coefficient of Determination
(r2)
0.979
0.341
0.999
0.943
0.926
0.988
Mean (± SD)
(%)
-31.3 (±9.5)
-5.6 (±22.1)
1.1 (±21.8)
74.8 (± 54.3)
-0.70 (±8.1)
-20.5 (± 15.4)
Hourly Data (%)
95.4 to 96.0
Median
(%)
-31.0
1.2
-0.60
61.9
-1.3
-17.9
Daily Averages
with > 12 Hours
(%)
96.9

            CI = Confidence interval; SD = Standard deviation
 Signed by
October 10. 2012
Tracy Stenner                      Date
Manager
Environmental Solutions Product Line
Energy, Environment, and Materials Sciences
Battelle
Signed by
October 31. 2012
                                     Cynthia Sonich-Mullin              Date
                                     Director
                                     National Risk Management Research Laboratory
                                     Office of Research and Development
                                     U.S. Environmental Protection Agency
          NOTICE: ETV verifications are based on an evaluation of technology performance under specific,
          predetermined criteria and the appropriate quality assurance procedures. EPA and Battelle make no
          expressed or implied warranties as to the performance of the technology and do not certify that a technology
          will always operate as verified. The end user is solely responsible for complying with any and all applicable
          federal, state, and local requirements. Mention of commercial product names does not imply endorsement.

-------