EPA/600/R-11/181
                                           www.epa.gov/narel
                                             November 2011
                                                 Revision 0
Rapid Radiochemical Method for Phosphorus-32
     in Water for Environmental Remediation
       Following Homeland Security Events
               U.S. Environmental Protection Agency

                   Office of Air and Radiation
                Office of Radiation and Indoor Air
          National Air and Radiation Environmental Laboratory
                    Montgomery, AL 36115

               Office of Research and Development
            National Homeland Security Research Center
                     Cincinnati, OH 45268

-------
                   Phosphorous-32 in Water: Rapid Method for High Activity Samples
                                     Revision History
Revision 0   | Original release.	|	11/17/2011
 This report was prepared for the National Air and Radiation Environmental Laboratory of the Office of
 Radiation and Indoor Air and the National Homeland Security Research Center of the Office of Research
 and Development, United States Environmental Protection Agency. It was prepared by Environmental
 Management Support, Inc., of Silver Spring, Maryland, under contract EP-W-07-037, work assignment I-
 41, managed by David Carman  and Dan Askren. Mention of trade names or specific applications does
 not imply endorsement or acceptance by EPA.

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
      Rapid Radiochemical Method for Phosphorus-32 in Water for Environmental
                    Remediation Following Homeland Security Events
1.  Scope and Application
   1.1. The method will be applicable to water samples where radioactive contamination is
        either from known or unknown origins. If any filtration of the sample is performed
        prior to starting the analysis, those solids should be analyzed separately. The results
        from the analysis of these solids should be reported separately (as a suspended activity
        concentration for the water volume filtered), but identified with the filtrate results.
   1.2. The method is specific for 32P in drinking water and other aqueous samples. 33P, another
        long-lived isotope of phosphorous, may be present in the sample. However, since the
        beta particle energy is insufficient to generate Cerenkov radiation,  it will not interfere
        with analysis for P by this method.
   1.3. The method uses rapid radiochemical separation techniques for determining 32P in water
        samples following a radiological or nuclear incident. Although the method can detect
        concentrations of 32P on the same order of magnitude as methods used for the
        measurement of gross beta concentration for the Safe Drinking Water Act (SDWA), the
        method cannot be used for SDWA applications because no 32P-specific method has
        been approved by EPA.
   1.4. The method is capable of achieving a relative required method uncertainty for 32P of
        13% at an analytical action level (corresponding to a 100-mrem dose rate for 50 years)
        of 12,000 pCi/L with 100 mL of sample and a counting time less than 30 minutes (see
        Steps 9.2 and 9.3). A larger sample size, followed by concentration and purification of
        phosphate, and a longer counting time is necessary to achieve a comparable SDWA
        detection level for gross beta concentration of 3 pCi/L (see Step  9.4). The Safe
        Drinking Water Act Maximum Contaminant Level (MCL) for 32P is 30 pCi/L.
   1.5. The sample turnaround time and throughput may vary based on additional  project
        MQOs, the time for analysis of the final sample test source,  and initial sample volume.
        The method must be validated prior to use following the protocols provided in Method
        Validation Guide for Qualifying Methods  Used by Radiological Laboratories
        Participating in Incident Response Activities (See Reference 16.1).
   1.6. The method is intended to be used for water samples that are similar in composition to
        drinking water or surface water. The rapid 32P method was evaluated following the
        guidance presented for "Level E Method Validation: Adapted or Newly Developed
        Methods, Including Rapid Methods" in Method Validation Guide for Qualifying
        Methods Used by Radiological Laboratories Participating in Incident Response
        Activities (see Reference 16.1) and Chapter 6 of Multi-Agency Radiological Laboratory
        Analytical Protocols Manual (MARLAP 2004, Reference 16.2). The matrix used for
        the determination of 32P was drinking water from Atlanta, GA.
                                                          ^9
   1.7. The method is applicable to the determination  of soluble  P. The method is not
        applicable to the determination of 32P in highly insoluble particulate matter possibly
        present in water samples contaminated as  a result of a radiological dispersion device
        (RDD) event. The source of such materials may be brachytherapy sources encased in a
        Ti-Ni casing. The proprietary nature of the chemical form of these sources makes it
        difficult to establish a chemical analysis method. Suffice it to say that the acid digestion
11/17/2011                                    1                              Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
         steps used in this method would be insufficient to dissolve such sources. This also
         means that they would not be available for intake and accumulation in the body. For
         this reason, the water sample is filtered first. The filtered residue should be retained for
         subsequent analysis.

2.   Summary of Method
    2.1. A 100 mL water sample is filtered and phosphate carrier is added. The solution is passed
         through a cation exchange resin and then a Diphonix® resin to remove interferences
         from cation radionuclides. The eluent is treated with a mixture of 10 mL of 30 % H2O2
         and 10 mL of concentrated nitric acid, reduced in volume, by heating, to approximately
         2 to 5 mL, and quantitatively transferred to a LSC vial for counting. The Cerenkov
         photons from the 32P beta (1710 keV, Emax) decay are detected using a calibrated liquid
         scintillation counter (LSC). Following counting of the sample, an aliquant of the final
         solution is used for yield determination by the inductively coupled plasma-atomic
         emission spectrometry (ICP-AES) method.

3.   Definitions, Abbreviations, and Acronyms
    3.1. Analytical Protocol Specifications (APS). The output of a directed planning process that
         contains the project's analytical data needs and requirements in an organized, concise
         form.
    3.2. Analytical Action Level (AAL). The term "analytical action level" is used to denote the
         value of a quantity that will cause the decisionmaker to choose one of the alternative
         actions.
    3.3. Analytical Decision Level (ADL). The analytical decision level refers to the value that is
         less than the AAL and is based on the acceptable error rate and the required method
         uncertainty.
    3.4. Discrete Radioactive Particles (DRPs or Hot Particles). Paniculate matter in a sample of
         any matrix where a high concentration of radioactive material is contained in a tiny
         particle (diameter is in the micro-meter range).
    3.5. Figure of Merit (FOM). The figure of merit is a measure of the response  to the analyte
         by the instrument relative to the background. The FOM is the square of the detector
         efficiency divided by the background count rate, both measured in the same region of
         the spectrum.
    3.6. Liquid Scintillation Counter or Counting (LSC). A beta spectrometer used to measure
         Cerenkov radiations.
    3.7. Multi-Agency Radiological Laboratory Analytical Protocols Manual (see Reference
         16.2).
    3.8. Measurement Quality Objective (MQO). MQOs are the analytical data requirements of
         the data quality objectives and are either project or program specific. They can be
         quantitative or qualitative. MQOs serve as measurement performance criteria or
         objectives of the analytical process.
    3.9. Radiological Dispersal Device (ROD), i.e., a "dirty bomb." This is an unconventional
         weapon constructed to distribute radioactive material(s) into the environment either by
         incorporating them  into a conventional bomb or by using sprays, canisters, or manual
         dispersal.
11/17/2011                                     2                              Revision 0

-------
                   Phosphorous-32 in Water: Rapid Method for High Activity Samples
   3.10. Region of interest (ROI). The region of interest is that part of the spectrum for a
         particular analyte or group of analytes where it is expected that the instrument will have
         the most analytically favorable response. The ROI is user selected, often by using the
         FOM calculation for different regions of the spectrum.
   3.11. Required Method Uncertainty (MMR). The required method uncertainty is a target value
         for the individual measurement uncertainties and is an estimate of uncertainty (of
         measurement) before the sample is actually measured. The required method uncertainty
         is applicable at or below an AAL.
   3.12. Required Relative Method Uncertainty ($JMR). The required relative method uncertainty
         is the MMR divided by the AAL and typically expressed as a percentage. It is applicable
         above the AAL.
   3.13. Sample Test Source (STS). This is the final form of the sample that is used for nuclear
         counting. This form is usually specific for the nuclear counting technique used in the
         method, such as a solid deposited on a filter for alpha spectrometry analysis.

4.  Interferences
   4.1. Radiological: Cerenkov counting measures photons generated by high energy beta
         particles passing through the final test solution regardless of the radionuclide. The
         threshold beta particle energy that can produce Cerenkov radiation with a counting
                                      1 	                               "29
         efficiency of-6% is ~ 900 keV. Therefore, radionuclides other than  P in the final
         test sample solution having maximum beta emissions greater than 900 keV would
         likely cause measurable interferences and bias results high. Below 900 keV, the
         efficiency for Cerenkov counting decreases significantly. Table 17.1 lists some other
         radionuclides that have sufficient high-energy beta emissions to be counted efficiently
         by Cerenkov counting. This method provides a technique for the chemical  purification
         of phosphorous from these other potential radionuclides (e.g., fission products, 40K and
         decay products of the natural occurring series) having sufficient beta energies to
         produce Cerenkov radiations. Although it is  unlikely that these radionuclides would
         accompany 32P in a ROD, it is important that the method discriminate against them in
         the final sample test source. Radiological interferences in the final sample test source
         can be evaluated by counting the test source solution approximately  7 days from the
         first count, and also by examining the energy distribution in the liquid scintillation
         spectrum. The second count should produce a net count rate that is ~ 71% of the count
         rate of the first test source measurement. Inferences also may be detected by examining
         the energy distribution in the individual sample spectra and comparing it to the
         expected, interference-free spectrum.
   4.2. Non-Radiological: High  levels of phosphates  (>0.1 mg/L) in the water are accounted for
         by the relative yielding technique used in this rapid method. If suspended solids are
         observed or known to exist in the sample, the sample should be filtered through a 0.45-
         um filter before proceeding with the method.

5.  Safety
   5.1. General
1 See Section 4.5.1 of NCRP Report No. 58, reference 16.7). A state-of-the art liquid scintillation counter may have
a better response to the Cerenkov radiation and provide higher detector efficiencies than those stated in the
reference.


11/17/2011                                      3                              Revision 0

-------
                   Phosphorous-32 in Water: Rapid Method for High Activity Samples
         5.1.1.  Refer to your safety manual for concerns of contamination control, personal
                exposure monitoring, and radiation dose monitoring.
         5.1.2.  Refer to the laboratory chemical hygiene plan (or equivalent) for general safety
                rules regarding chemicals in the workplace.
   5.2. Radiological
         5.2.1.  Hot Particles (DRPs)
                5.2.1.1.  Hot particles will be small, on the order of 1 mm or less. Typically,
                        DRPs are not evenly distributed in the media and their radiation
                        emissions are not uniform in all directions (anisotropic). Filtration
                        using a 0.45-um (or finer) filter will minimize the presence of these
                        particles.
                5.2.1.2.  Care should be taken to provide suitable containment for filter media
                        used in the pretreatment of samples that may have DRPs because the
                        particles become highly statically charged as they dry out and will
                        "jump" to other surfaces, causing contamination.
                5.2.1.3.  Filter media should be individually surveyed for the presence of these
                        particles and this information should be reported with the final sample
                        results.
         5.2.2.  For samples  with detectable activity concentrations of this radionuclide, labware
                should be used only once due to potential for cross contamination.
   5.3. Procedure-Specific Non-Radiological Hazards
         • None

6.  Equipment and Supplies
   6.1. Analytical balance with a 0.001-g readability or better.
   6.2. Toploader balance with a 0.1-g readability
   6.3. Beakers, Pyrex® - 250, 400 mL
   6.4. Hot plate, or other suitable device for reducing sample volume
   6.5. Glass  stirring rods
   6.6. Graduated cylinders - 25, 50, 100, 250, 1,000 mL
   6.7. Pipettes, volumetric / automatic: assorted volumes down to the microliter range.
   6.8. Scintillation vials - 22 mL glass
        NOTE: It has been demonstrated that plastic vials yield a higher efficiency for Cerenkov counting
        than do glass vials.
   6.9. Volumetric flasks - 25, 100, 200, 500, 1,000 mL.
   6.10. Detector capable of  measuring Cerenkov radiation - Liquid scintillation counter
   6.11. Inductively Coupled Plasma-Atomic Emission Spectrometer  (ICP-AES)

7.  Reagents and Standards
   NOTES: All reagents are American Chemical Society (ACS) reagent grade or equivalent unless otherwise
          specified.
          Unless otherwise indicated, all references to laboratory water should be understood to mean
          ASTM D1193 Type I Reagent water (see Reference 16.3).
   7.1. Cation exchange resin: DOWEX 50WX8  50-100 mesh,  lOmL
   7.2. Diphonix® resin (Eichrom): 100-200 mesh, 2.0 mL
   7.3. Hydrogen peroxide (H2O2): (30%)
   7.4. Demineralized water  (DM)
11/17/2011                                      4                               Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
   7.5. Carrier solution: Commercially available TCP standard for phosphorus diluted tol.OO
        mg/mL P, or an approximate 10 mg/mL P solution prepared from Na3PO4»12H2O
        standardized using the ICP-AES and diluted to 1.00 mg/mL P.
   7.6. Nitric acid (HNOs): concentrated (16 M). Note: HNOs can degrade forming colored
        contaminants (nitric oxides) which will interfere with Cerenkov counting. A new bottle
        of HNOs free of color contaminants should be used.
        7.6.1. Nitric acid (HNO3): 2 M; dilute 250 mL of concentrated HNO3 to 500 mL with
              DM. Cool and dilute to 1,000 mL with DM.

8.  Sample Collection, Preservation, and Storage
   8.1. Collect a 1 liter sample in the field in a suitable container.
   8.2. No sample preservation is required if sample is delivered to the laboratory within 3 days
        of sampling date/time.
   8.3. If the dissolved concentration of 32P is sought, the insoluble fraction must be removed by
        filtration before preserving with acid.
   8.4. If the sample is to be held for more than 3 days, concentrated HNOs shall be added to
        achieve a pH < 2.

9.  Quality Control
   9.1. Batch quality control results shall be evaluated and meet applicable analytical protocol
        specifications (APS) prior to release of unqualified data. In the absence of project-
        defined APS or a project-specific quality assurance project plan (QAPP), the quality
        control  sample acceptance criteria defined in the laboratory quality manual and
        procedures shall be used to determine acceptable performance for this method.
        9.1.1.  A laboratory control sample (LCS) shall be run with each batch of samples. The
               concentration of the LCS shall be at or near the action level or level of interest
               for the project.
        9.1.2.  One method blank shall be run with each batch of samples.  The laboratory
               blank should consist of laboratory water.
        9.1.3.  One laboratory duplicate shall be run with each batch of samples. The
               laboratory duplicate is prepared by removing an aliquant from the  original
               sample  container.
        9.1.4.  A matrix spike sample may be included as a batch quality control sample if
               there is concern that matrix interferences may compromise chemical yield
               measurements or overall data quality.
   9.2. This method is capable of achieving a WMR of 150 pCi/L at or below an action level of
        12,000 pCi/L.
   9.3. This method is capable of achieving a ^MR 13% above 12,000 pCi/L. This may be
        adjusted if the event specific MQOs are different.
   9.4. This method is capable of achieving a required minimum detectable concentration
        (MDC) of 3 pCi/L, provided that a minimum sample size of 500 mL is used and a count
        time of at least 100 min is performed.

10. Calibration and Standardization
   10.1. Initially set up the liquid scintillation spectrometer according to the manufacturer's
        recommendations. The energy range of the spectrometry system should be capable of
11/17/2011                                     5                              Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
        measuring in the region of interest (ROI) between 30 and 2,000 keV. Go to Step 10.3 to
        establish the 32P ROI energy window that will be used to analyze 15-mL samples of a
        2-M HNOs sample test source.
    10.2. Conduct a final set up of the Liquid Scintillation Spectrometer for use to count samples
        according to ASTM Standard Practice D7282, Section 9.4, "Liquid Scintillation
        Counting Initial Instrument Set-up" (see Reference 16.5). Continuing Instrument
        Quality Control Testing shall be performed according to ASTM  Standard Practice
        D7282, Section 25, Quality Control for the Liquid Scintillation Counter
    10.3. Liquid Scintillation Counter Region-of-Interest  Setup
        10.3.1.   The normal photomultiplier high voltage and amplifier gain settings used for
                 liquid scintillation counting (LSC) are satisfactory for Cerenkov counting. In
                 modern LS counters, the beta/Cerenkov radiation interactions in the test
                 sample generate photons detected by photomultiplier tubes  (PMT). The PMT
                 in turn converts the detected photon energy voltage pulse which is amplified
                 and transferred to an analysis component of the counter. The voltage pulses
                 are separated by voltage height and digitally converted  into an energy
                 spectrum through the use of a multichannel analyzer. The resulting energy
                 spectrum is divided into energy channels; the number of energy channels
                 covers the maximum beta particle or Cerenkov energy of interest (e.g., 4000
                 channels for 0 to 2,000 keV). For most applications, a ROI  of the spectrum is
                 selected for a given radionuclide. Because of the relatively lower number of
                 photons generated and the lower photon-to-electron generation ratio compared
                 to LSC, the maximum beta energy needed for most practical Cerenkov
                 counting is about 1,000 keV, although radionuclides having lower maximum
                 beta energies can be measured.2 The Cerenkov counting efficiency for 32P
                 (Pmax = 1710 keV) is expected to be ~ 55%.
        10.3.2.   In a similar manner for LSC instrument set up, the ROI to be used for the
                                                           17
                 qualitative and quantitative determination of a  P must be established prior to
                 instrument calibration and radionuclide quantification.  The figure of merit
                 (FOM; efficiency2/background) concept shall be used to determine the most
                 advantageous 32P ROI. Once established, the ROI must be maintained during
                                                                       on
                 the operation of instrument. A quality control chart using a   Sr/Y source
                 should be used to ensure the ROI counting efficiency of the instrument has not
                 changed and the 32P ROI has not shifted.
        10.3.3.   The same LSC setting, LS vial type, sample volume, and acidic concentration
                 must be maintained during the establishment of the ROI and instrument
                               17
                 calibrations for  P and instrument background.
        10.3.4.   The volume to the shoulder of the LSC vial is very reproducible. Using this as
                 a volumetric measure for the standards and samples saves time, sample
                 handling, and reduces potential for cross-contamination. Each laboratory
                 should measure this volume gravimetrically on their own. Typically, this
                 measurement shows a variability of less than 2%.
 See Section 4.5.1 of NCRP Report No. 58, Reference 16.7. Radionuclides with lower beta maximum energies can
be measured, but efficiencies are quite low.


11/17/2011                                     6                              Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
        10.3.5.   Set the lower level discriminator of the LSC instrument to eliminate excessive
                 noise signals that give rise to low-energy counts. This setting will vary
                 depending on the instrument manufacturer.
   10.4. 32P Region of Interest Determination
        10.4.1.   Count a 32P background vial containing 15.0 mL of 2 M HNO3 for 400
                 minutes. Label the vial and spectrum as 32PeKG and electronically save the
                 spectrum.
        10.4.2.   Use the 32P spectrum generated from the LSC Calibration (Step 10.3.2), set
                 the upper ROI channel to the end point channel of the 32P energy distribution,
                 32PEPC- Sum the number of counts from channel 10 to 32PEpc and divide by the
                                         19
                 counting time. Note this as  Pgcpmi •
        10.4.3.   Using the 32PeKG spectrum from the background sample vial, sum the number
                 of counts from channel 10 to 32PEPc and divide by the counting time. Note this
                   32n
                 as  -TBKGcpml
        10.4.4.   Calculate the relative FOM value as:
                               32P    _ 32p      f/32p
                                 gcpm;    1BKGcpm1/ /  1 BKGcpmj

                 Record this as the 32P mu value.
        10.4.5.   In a stepwise manner, gradually increase the lower channel included in the
                                                                      19            19
                 ROI and sum the counts in a new ROI (e.g., channel 20 to  PEPC) for the  P
                     19                      19        19            19
                 and  PBKG spectra. Calculate  P gcpni2,  PBKGcpm2 and  PpoM2- Continue
                                                                                    19
                 decreasing the ROI by incrementing the lower channel by 10 and calculate  P
                 cpmi; 32PBKGcPmi and 32PFOM; values.
                           19
        10.4.6.   Review the  PFOM values and select the ROI with the highest FOM value.
                                     19
                 Record the ROI as the P ROI and use this window when calibrating the unit
                    19
                 for  P detector efficiency as provided in Equation 1 .
   10.5. 32P Region-of-Interest Counting Efficiency
        10.5.1.   Using the 32PBKG spectrum obtained in Step 10.4.1, calculate the background
                 count rate in the 32PRoi.
        10.5.2.   Prepare a calibration source by adding an appropriate amount of traceable 32P
                 concentration (at least 1,000 dpm) in a 5 mL solution of 8 M HNOs. Bring the
                 final solution volume to 15.0 mL with DM water. Note the reference time and
                 date of preparation and reference concentration (dpm) in the sample.
        10.5.3.   Count the  2P vial to obtain at least 10,000 counts in the same ROI energy
                 window optimized for 32P as determined in Step 10.4.6.
        10.5.4.   Calculate the 32P ROI fractional detection efficiency using the following
                 equations:
                                   e__(R,-R^DC
                                        ACxVxDF                                   ( '
                 where
                                    DF  = Q-*™^-V                                   3
11/17/2011                                     7                              Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
                                   DC=     p-?°                                    (4)
                                         (l-e-V-^))


                 s     =  detection efficiency for 32P
                                19
                 Rs    =  gross  P ROI count rate (cpm) for the working calibration source
                 Rb    =  count rate (cpm) of the background subtraction sample in the 32P
                          ROI
                 AC   =  concentration (dpm/volume or mass) of the 32P standard solution as
                          of its reference date
                 V    =  amount (volume or mass) of the standard solution added
                 DF   =  correction factor for decay of the standard from its reference date
                          through the measurement of the vial
                 DC   =  correction factor for decay during counting
                 Ap.32  =  decay constant for 32P, 3.375x 1CT5 min"1
                 to     =  reference date and time for the 32P standard
                 t\     =  date and time of the start of the Cerenkov counting measurement for
                              19
                          the  P standard
                 tc     =  counting time in minutes
11. Procedure
   11.1. Water Sample Preparation
        11.1.1.  As required, filter enough sample volume through a 0.45-um filter to provide
                 at least 200 mL for radiochemical analysis and sufficient volume for ICP-AES
                 analysis.
                 NOTES: A volume of at least 500 mL will be necessary to achieve a detection limit of 3
                        pCi/L. If the 3 pCi/L LLD is to be achieved in subsequent steps:
                        A sample volume of 500 mL should be used.
                        All additions of chemical volumes should be scaled up by a factor of 5.
        11.1.2.  Transfer 150 mL of the filtered sample to an appropriate container.
        11.1.3.  If not acidified in the field, add sufficient concentrated HNOs to the sample to
                 reach a pH of less than 2.0. This usually requires about 0.3 mL of HNOs per
                 150 mL of sample.
        11.1.4.  Add 1.0 mL of the 1.00 mg/mL phosphorous carrier solution to the beaker and
                 swirl to mix.
        11.1.5.  Use an aliquant of the sample to determine the phosphorus concentration in
                 |ig/mL by ICP-AES. Record the concentration as  Pconcr
        11.1.6.  Prepare a 10-mL cation exchange column in tandem with a 2 mL Diphonix®
                 column.
        11.1.7.  Pass about 20 mL of the sample solution through the tandem column
                 arrangement and discard the  eluate.
                 NOTES: This volume is sufficient to equilibrate the columns with the sample solution.
                        The flow rate through the columns may be controlled using devices such as a
                        vacuum box or peristaltic pump. The flow rate should not exceed about 5
                        mL/minute.
11/17/2011                                     8                               Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
         11.1.8.   Pass the remaining sample solution through the columns and collect the eluate
                 in an appropriate container.
         11.1.9.   Measure a known volume (approximately 100 mL) of the sample which has
                 passed through the columns into a 250 mL beaker for performing evaporation
                 and digestion on a hot plate.
         11.1.10.  Add 10 mL of 30% H2O2 and 10 mL of concentrated nitric acid.
         11.1.11.  Bring the solution to near boiling and reduce the volume to between 2 and 5
                 mL.
   1 1.2. Preparation of the sample test source
         1 1 .2. 1 .   Transfer the solution to a liquid scintillation vial and rinse the beaker three
                 times  with 5-mL aliquants  of 2-M nitric acid.
         1 1 .2.2.   Bring the level in the liquid scintillation vial to the shoulder of the vial (refer
                 to Step 10.3.4),  cap, and invert several times to mix.
         1 1.2.3.   Count the liquid scintillation vial for a period of time sufficient to achieve a
                 WMnof 150pCi/L.
         11.2.4.    After  counting the sample test source, remove an aliquant and determine the
                 phosphorous concentration (|ig/mL) by ICP-AES. Record the concentration as
12. Data Analysis and Calculations
   12.1. Equation for determination of final result, combined standard uncertainty, and chemical
        yield (if requested):

        The activity concentration of an analyte and its combined standard uncertainty are
        calculated using the following equations:

        12.1.1.   The 32P activity concentration (AC) is calculated as follows:
                    where:
                                AC=     .
                                      2.22xsxYxVxDF
                                                                                      (7)
                                         (l-e^p-32(y)

                    and where:
                       R&      =    gross count rate for the sample (cpm) in 32P ROI
                       Rb      =    background count rate (cpm) 32P ROI
                       s       =    efficiency of the detector for 32P ROI
                       7       =    fractional chemical yield for phosphorous
                       V       =    volume of the sample aliquant (L)
                       DF     =    correction factor for decay of the sample from its
                                     reference date until the start of the 32P count
                       DC     =    correction factor for decay during counting
                       Ap.32     =    decay constant for 32P, 3.375x 10~5 min"1
11/17/2011                                     9                               Revision 0

-------
                   Phosphorous-32 in Water: Rapid Method for High Activity Samples
                         to        =    reference date and time for the sample
                         t\        =    date and start time of counting the sample
                                       Note: the differential time t\ - to must be in minutes
                         tc        =    counting time in minutes

                  Note: The elapsed time between the sample count and the reference date must be
                  calculated using the same time units as the decay constant

         12. 1 .2.   The standard counting uncertainty of the phosphorous activity concentration,
                  wcc, is calculated as follows:
                                             *.    tb
                                  cc   2.22 xsxYxVxDF
                  where:
                     4   =  duration of the sample count (min), and
                     tb   =  duration of the background subtraction count (min).


         12.1.3.   The combined standard uncertainty (CSU) for the 32P activity concentration,
                  uc(ACp-32)-, is calculated as follows:

                     uc(ACP.32) = Ju2c + AC2 x  !—^- + -
                                  \            I  £      1       y   J
              where:
              w(Y)       =  standard uncertainty of fractional chemical yield for phosphorous,
              z/(V)       =  standard uncertainty of the volume of the sample aliquant (L), and
              u(s)       =  standard uncertainty of the 32P ROI detector efficiency, s.
         12.1.4.   If the critical level concentration (Sc) or the minimum detectable
                  concentration (MDC) are requested (at an error rate of 5%), they can be
                  calculated using the following equations:3
              0.4x   --1 +0.677x 1 +  -  +1.645x
                                     t x2.22xsxYxVxDF
:DC

	     (10)
3 The formulations for the critical level and minimum detectable concentration are based on the Stapleton
Approximation as recommended in MARLAP Sections 20A.2.2, Equation 20.54, and 20A.3.2, Equation 20.74,
respectively. The formulations presented assume a = 0.05, /? = 0.05 (with zi_a = Zj.p = 1.645), and d = 0.4.


11/17/2011                                       10                                Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
                   MDC =
2.71x1 1 + -?- | + 3.29x |^s x  1 + -
                                                                 :DC
                                     t x2.22xsxYxVxDF
                                                          (11)
   12.2. Chemical Yield for Phosphorous
        12.2.1.   Calculate the chemical yield for phosphorous using the information gathered
                 in Steps 11.1.5 and 11.2.4.

                 The chemical yield of the carrier (7p) and its standard uncertainty u(Yp) can
                 be calculated using the following equations:

                                       YP = y-                                       (12)

                 where:

                   PI    =    (Pconcj in |ig/mL - refer to Step 11.1.5) x (sample volume used
                              in procedure in mL - refer to Step 11.1.9)

                              NOTE:  Pt is the mass of P (jig) prior to performing procedure

                   PY   =    (Pconc2 in |ig/mL - refer to Step 11.2.4) x (volume of solution in
                              LCS vial in mL - refer to Step 10.3.4)

                              NOTES:  Pt is the mass of P (fig) added as carrier
                                     PY is the mass of P (fig) recovered after completing procedure.

                 and:

                                     w(7p)~0.02                                   (is)

   Because the yield determination is a relative comparison using the same ICP-AES
   instrument, the uncertainty budget is related to the precision of the measurements, which is
   estimated at 2% (assumes that all volumetric dilutions are much less that 2% and thus
   contribute negligibly).

   12.3. Results Reporting
        12.3.1. The following items should be reported for each result: sample ID, volume of
               sample used, carrier yield and its uncertainty, and Cerenkov counting efficiency
               and its uncertainty.
        12.3.2. The following items should be reported for each result:
               12.3.2.1. Result in scientific notation ± combined standard uncertainty, critical
                        level, and MDC.
               12.3.2.2. If solid material was filtered from the solution and analyzed
                        separately, the results of that analysis should be reported separately as
                        pCi/L of the original volume from which the solids were filtered if no
11/17/2011                                     11                               Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
                       other guidance is provided on reporting of results for the solids. For
                       example:
                                32P for Sample 12-1-99:
                                       Filtrate Result: 12.8 ± 1.5 pCi/L
                                       Filtered Residue Result: 2.5 ± 0.3 pCi/L

13. Method Performance
   13.1. Method validation results are to be reported as an attachment.
        13.1.1. Expected turnaround time per batch of 14 samples plus QC, assuming
               microprecipitations for the whole batch are performed simultaneously:
               13.1.1.1. For an analysis of a 100-mL sample aliquant, sample preparation, ion
                        exchange, digestion and evaporation should take ~ 5.5 h.
               13.1.1.2. Final test source sample preparation for Cerenkov counting takes
                        -Sminutes.
        13.1.2. A 100 minute counting time is sufficient to meet the MQOs listed in 9.2 and
               9.4, assuming detector efficiency of- 0.35 and chemical yield of at least 0.9.
               Longer counting time may be necessary to meet these MQOs if the yield is
               lower.
        13.1.3. Data should be ready for reduction between 8.5 and 10.5 h after beginning of
               analysis.

14. Pollution Prevention: This method utilizes a Diphonix® resin (-10 mL) and a cation resin
   (-10 mL). The removal of uranium, thorium, and TRU elements using the extraction column
   results in the elimination of solvent extraction and other multiple step precipitation and ion
   exchange techniques.

15. Waste  Management
   15.1. Types of waste  generated per sample analyzed
        15.1.1. Two mL of Diphonix® resin and 10 mL of cation exchange resins are generated
               per sample. If a TRU column is used to remove transuranics, this is -2 mL.
        15.1.2. Approximately 20 mL of acidic waste from the final sample tests source are
               generated
        15.1.3. Unless processed further, the TRU resin may contain isotopes of uranium,
               neptunium, and thorium, if any of these were present in the sample originally.
   15.2. Evaluate all waste streams according to disposal requirements by applicable
        regulations.

16. References
   16.1.   U.S. Environmental  Protection Agency (EPA). 2009. Method Validation Guide for
          Qualifying Methods  Used by Radiological Laboratories Participating in Incident
          Response Activities.  Revision 0. Office of Air and Radiation, Washington, DC. EPA
          402-R-09-006, June. Available at: www.epa.gov/narel/incident_guides.html and
          www.epa.gov/erln/radiation.html.
   16.2.   Multi-Agency Radiological Laboratory Analytical Protocols Manual (MARLAP).
          2004. EPA 402-B-1304 04-001 A, July. Volume I, Chapters 6, 7, 20, Glossary;
11/17/2011                                    12                              Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
          Volume II and Volume III, Appendix G. Available at:
          www.epa.gov/radiation/marlap/links.html.
   16.3.   ASTM Dl 193, "Standard Specification for Reagent Water" ASTM Book of
          Standards  11.01, current version, ASTM International, West Conshohocken, PA.
          Available at: www.astm.org/Standards/Dl 193.htm.
   16.4.   National Council on Radiation Protection and Measurements (NCRP). 1985. Report
          No. 58, A Handbook of Radioactivity Measurements Procedures., Second Edition.
          Available from www.ncrponline.org/Publications/Publications.html.
   16.5.   ASTM D7282 "Standard Practice for Set-up, Calibration, and Quality Control of
          Instruments Used for Radioactivity Measurements," ASTM Book of Standards 11.02,
          current version, ASTM International, West Conshohocken, PA. Available at:
          www.astm.org/Standards/D7282.htm.
   16.6.   Colle, R. 1999. "Chemical Digestion and Radionuclidic Assay of TiNi-Encapsulated
          32P Intravascular Brachytherapy Sources" Applied Radiation and Isotopes,  50, 811-
          833.
11/17/2011                                    13                              Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
17. Tables, Diagrams, Flow Charts, and Validation Data
    17.1.  Table 17.1, Beta Particle Energy and Abundance
          Including major radiation emissions from all radionuclides separated.
Radionuclide
32p
33p
Half-Life
(days)
14.262
25.34
>,
(min"1)
3.374xlO~5
1.899xlO~5
Abundance
1.00
1.00
Energy (keV)
Maximum
Average
1710
695
249
76.4
Potential Radiological Interferences
90Sr(90Y)
89Sr
106Ru (106Rh)
140Ba(140La)
4oK
137Cs
124Sb
110mAg
l.OSxlO4
50.5
373
12.75
4.55x10"
1.103xl04
60.2
250
4.58xlO~8
9.53xlO~6
1.29xlO~6
3.77xlO~5
1.06xlO~15
4.38xlO~8
8.0x10^
1.93xlO~6
1.00
1.00
0.79
0.20
0.89
0.053
0.232
0.0129
2280
934
1495
584
3541
1508
1679
630
1311
560
1176
416
2302
918
2892
1199
11/17/2011
14
Revision 0

-------
                    Phosphorous-32 in Water: Rapid Method for High Activity Samples
    17.2.  Spectra
          17.2.1.  Spectrum4 of Mixture of 32P and 33P
                                                                      32p
                     °   2345    1Q1    2343   1Q2   £341    \ Q3    2

                                             E/keV

          17.2.2.  Spectrum of Processed Sample
                                           Total Counts vs. Channel
        3000
        2500
        20CO
      » 1500 ffl

      I
      o
        1000
         500
                                   :15-'. ,:  I 20
                                                          :.30- '
        -500
                                                 Channel #
4 From "Chemical Digestion and Radionuclidic Assay of TiNi-Encapsulated 32P Intravascular Brachytherapy
Sources," Reference 16.10.
11/17/2011
15
Revision 0

-------
                  Phosphorous-32 in Water: Rapid Method for High Activity Samples
   17.3. Decay Scheme
        100 %
        Ep max = 249 keV
                     100 %
                     ER max= 1710 keV
                                             32S
                                           Stable
11/17/2011
16
Revision 0

-------
               Validation of Rapid Radiochemical Method for Americium-241 on Air Filters
    17.4. Flow Chart for Separation with Timeline
                         Sample preparation (Steps 11.1.1-11.1.3)
                        Filter, acidify, and aliquant for sample and native
                                   phosphorous analysis
                                        (
-------