EPA Document #: EPA/600/R-13/119
METHOD 540.   DETERMINATION OF SELECTED ORGANIC CHEMICALS IN
               DRINKING WATER BY SOLID PHASE EXTRACTION AND
               LIQUID CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY
               (LC/MS/MS)
                                Version 1.0
                              September 2013
J.A. Shoemaker   US EPA, Office of Research and Development, National Exposure
                Research Laboratory

D.R. Tettenhorst  US EPA, Office of Research and Development, National Exposure
                Research Laboratory
              NATIONAL EXPOSURE RESEARCH LABORATORY
                OFFICE OF RESEARCH AND DEVELOPMENT
               U. S. ENVIRONMENTAL PROTECTION AGENCY
                         CINCINNATI, OHIO 45268
                                   540-1

-------
                                   METHOD 540

           DETERMINATION OF SELECTED ORGANIC CHEMICALS IN
        DRINKING WATER BY SOLID PHASE EXTRACTION AND LIQUID
       CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY (LC/MS/MS)
1.  SCOPE AND APPLICATION
   1.1
   1.2
   1.3
This is a liquid chromatography/tandem mass spectrometry (LC/MS/MS) method for
determination of organic contaminants in drinking water.  Accuracy and precision data
have been generated in reagent water, and finished ground and surface waters for
compounds listed in the table below.
Analyte
3 -Hydroxycarbofuran
Bensulide
Chlorpyrifos oxon
Disulfoton sulfoxide
Fenamiphos
Fenamiphos sulfone
Fenamiphos sulfoxide
Methomyl
Phorate sulfone
Phorate sulfoxide
Tebuconazole
Tebufenozide
Chemical Abstract Services
Registry Number (CASRN)
16655-82-6
741-58-2
5598-15-2
2497-07-6
22224-92-6
31972-44-8
31972-43-7
16752-77-5
2588-04-7
2588-03-6
107534-96-3
112410-23-8
The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets
Data Quality Objectives (DQOs) that are developed based on the intended use of this
method. The single laboratory lowest concentration MRL (LCMRL) is the lowest true
concentration for which the future recovery is predicted to fall, with high confidence
(99%), between 50 and 150% recovery. Single laboratory LCMRLs for analytes in
this method range from 0.30-2.7 ng/L, and are listed in Table 5. The procedure used
to determine the LCMRL is described elsewhere.1

Laboratories using this method will not be required to determine the LCMRL for this
method, but will need to demonstrate that their laboratory MRL meets the require-
ments described in Section 9.2.4.
                                        540-2

-------
    1.4   Determining the Detection Limit (DL) for analytes in this method is optional (Sect.
         9.2.6). Detection limit is defined as the statistically calculated minimum concentration
         that can be measured with 99% confidence that the reported value is greater than zero.2
         The DL is compound dependent and is dependent on extraction efficiency, sample
         matrix, fortification concentration, and instrument performance. DLs for analytes in
         this method range from 0.15-1.1 ng/L, and are listed in Table 5.

    1.5   This method is intended for use by analysts skilled in solid phase extractions,
         operation of LC/MS/MS instruments, and the interpretation of associated data.

    1.6   METHOD FLEXIBILITY - In recognition of technological advances in analytical
         systems and techniques, the laboratory is permitted to  modify the evaporation
         technique,  separation technique, LC column, mobile phase composition, LC conditions
         and MS and MS/MS conditions (Sect. 6.9, 9.1.1, 10.2, and 12.1). Changes may not be
         made to sample collection and preservation (Sect. 8), sample extraction steps (Sect.
         11), or to quality control requirements (Sect. 9).  Method modifications should be
         considered only to improve method performance. Modifications that are introduced in
         the interest of reducing cost or sample processing time, but result in poorer method
         performance, should not be used. Analytes must be adequately resolved
         chromatographically in order to permit the mass spectrometer to dwell on a minimum
         number of compounds eluting within a retention time window.  Instrumental
         sensitivity  (or signal-to-noise) will decrease if too many  compounds are permitted to
         elute within a retention time window. In all cases where method modifications are
         proposed, the analyst must perform the procedures outlined in the initial demonstration
         of capability (IDC, Sect. 9.2), verify that all Quality Control (QC) acceptance criteria
         (Sect. 9) are met,  and that acceptable method performance can be verified in a real
         sample matrix (Sect. 9.3.6).

         NOTE: The above method flexibility section is intended as an abbreviated  summation
                 of method flexibility. Sections 4-12 provide detailed information of specific
                 portions of the method that may be modified.  If there is any perceived
                 conflict between the general method flexibility statement in Section 1.6 and
                 specific information in Sections 4-12, Sections 4-12 supersede Section 1.6.

2.  SUMMARY OF METHOD

    A 250-mL water sample is fortified with surrogates and passed through a solid phase
    extraction (SPE) cartridge to extract the method analytes and surrogates.  Compounds are
    eluted from the solid phase with a small amount of methanol. The extract is concentrated by
    evaporation with  nitrogen in a heated water bath, and then adjusted to a 1-mL volume with
    methanol  after adding the internal standards. A 10-|iL injection is made into an LC equipped
    with a Cig column that is interfaced to an MS/MS.  Analytes  are separated and identified by
    comparing the acquired mass spectra and retention times to reference spectra and retention
    times for calibration standards acquired under identical LC/MS/MS conditions. The
    concentration of each analyte is determined by using the internal standard technique.
                                          540-3

-------
   Surrogate analytes are added to all Field and QC Samples to monitor the extraction efficiency
   of method analytes.

3.  DEFINITIONS

   3.1   ANALYSIS BATCH - A set of samples that is analyzed on the same instrument
         during a 24-hour period, including no more than 20 Field Samples, that begins and
         ends with the analysis of the appropriate Continuing Calibration Check (CCC)
         standards.  Additional CCCs may be required depending on the length of the analysis
         batch and/or the number of Field Samples.

   3.2   CALTERATION STANDARD (CAL) - A solution prepared from the primary dilution
         standard solution and/or stock standard solution, internal standard(s), and the
         surrogate(s).  The CAL solutions are used to calibrate the instrument response with
         respect to analyte concentration.

   3.3   COLLISIONALLY ACTIVATED DISSOCIATION (CAD) - The process of
         converting the translational energy of the precursor ion into internal energy by
         collisions with neutral gas molecules  to bring about dissociation into product ions.

   3.4   CONTINUING CALIBRATION CHECK (CCC) - A  calibration standard containing
         the method analytes, internal  standard(s) and surrogate(s).  The CCC is analyzed
         periodically to verify the accuracy of the existing calibration for those analytes.

   3.5   DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be
         identified, measured, and reported with 99% confidence that the analyte concentration
         is greater than zero. This is a statistical determination  of precision (Sect. 9.2.6), and
         accurate quantitation is not expected at this level.2

   3.6   EXTRACTION BATCH - A set of up to 20 Field Samples (not including QC
         samples) extracted together by the same person(s) during a work day using the same
         lot of SPE devices,  solvents, surrogate, internal standard and fortifying solutions.
         Required QC  samples include Laboratory Reagent Blank, Laboratory Fortified Blank,
         Laboratory  Fortified Sample Matrix, and either a Field Duplicate or Laboratory
         Fortified Sample Matrix Duplicate.

   3.7   FIELD DUPLICATES (FD1  and FD2) - Two separate samples collected at the same
         time and place under identical circumstances, and treated exactly the same throughout
         field and laboratory procedures. Analyses of FD1 and FD2 give a measure of the
         precision associated with sample collection, preservation, and storage, as well as
         laboratory procedures.

   3.8   INTERNAL STANDARD (IS) - A pure chemical added to an extract or standard
         solution in a known amount(s) and used to measure the relative response of other
         method analytes and surrogates that are components of the  same solution. The internal
                                         540-4

-------
      standard must be a chemical that is structurally similar to the method analytes, has no
      potential to be present in water samples, and is not a method analyte.

3.9   LABORATORY FORTIFIED BLANK (LFB) - A volume of reagent water or other
      blank matrix to which known quantities of the method analytes and all the preservation
      compounds are added in the laboratory. The LFB is analyzed exactly like a sample,
      and its purpose is to determine whether the methodology is in control, and whether the
      laboratory is capable of making accurate and precise measurements.

3.10  LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - A preserved field
      sample to which known quantities of the method analytes are added in the laboratory.
      The LFSM is processed and analyzed exactly like a sample, and its purpose is to
      determine whether the sample matrix contributes bias to the analytical results. The
      background  concentrations of the analytes in the sample matrix must be determined in
      a separate sample extraction and the measured values  in the LFSM corrected for
      background  concentrations.

3.11  LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A
      duplicate of the Field Sample used to prepare the LFSM. The LFSMD is fortified,
      extracted, and analyzed identically to the LFSM. The LFSMD is used instead of the
      Field Duplicate to assess method precision when the occurrence of method analytes is
      low.

3.12  LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other
      blank matrix that is treated exactly as a sample including exposure to all glassware,
      equipment, solvents and reagents,  sample preservatives, internal standards, and
      surrogates that are used in the analysis batch.  The LRB is used to determine if method
      analytes or other interferences are present in the laboratory environment, the reagents,
      or the apparatus.

3.13  LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The
      single laboratory LCMRL is the lowest true concentration for which a future recovery
      is expected,  with 99% confidence, to be between 50 and 150% recovery.1

3.14  MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by
      vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and
      reactivity data including storage, spill, and handling precautions.

3.15  MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
      reported as a quantitated value for a method analyte in a sample following analysis.
      This defined concentration can be no lower than the concentration of the lowest
      calibration standard for that analyte and can only be used if acceptable QC criteria for
      this standard are met. A procedure for verifying a laboratory's MRL is provided in
      Section 9.2.4.
                                     540-5

-------
   3.16  PRECURSOR ION - For the purpose of this method, the precursor ion is the
         protonated molecule ([M+H]+) of the method analyte. In MS/MS, the precursor ion is
         mass selected and fragmented by CAD to produce distinctive product ions of smaller
         m/z.

   3.17  PRIMARY DILUTION STANDARD (PDS) SOLUTION - A solution containing the
         analytes prepared in the laboratory from stock standard solutions and diluted as needed
         to prepare calibration solutions and other needed analyte solutions.

   3.18  PRODUCT ION - For the purpose of this method, a product ion is one of the fragment
         ions produced in MS/MS by CAD of the precursor ion.

   3.19  QUALITY CONTROL SAMPLE (QC S) - A solution of method analytes of known
         concentrations that is obtained from a source external to the laboratory and different
         from the source of calibration standards. The second source stock standard solution is
         used to fortify the QCS at a known concentration.  The QCS is used to check
         calibration standard integrity.

   3.20  STOCK STANDARD  SOLUTION (SSS) - A concentrated solution containing one or
         more method analytes prepared in the laboratory using assayed reference materials or
         purchased from a reputable commercial source.

   3.21  SURROGATE ANALYTE (SUR) - A pure chemical which chemically resembles
         method analytes and is extremely unlikely to be found in any sample. This chemical is
         added to a sample aliquot in known amount(s) before processing and is measured with
         the same procedures used to measure other method analytes. The purpose of the SUR
         is to monitor method performance with each sample.

4.  INTERFERENCES

   4.1   All glassware must be meticulously cleaned. Wash glassware with detergent and tap
         water, rinse with tap water, followed by a reagent water rinse.  Non-volumetric
         glassware can be heated in a muffle furnace at 400 °C for two hours or solvent rinsed.
         Volumetric glassware should be solvent rinsed and not be heated in an oven above
         120 °C.

   4.2   Method interferences may be caused by contaminants in solvents, reagents (including
         reagent water), sample bottles and caps, and other sample processing hardware that
         lead to discrete artifacts and/or elevated baselines in chromatograms. All items must
         be routinely demonstrated to be free from interferences (less than 1/3 the MRL for
         each method analyte) under the conditions of the analysis by analyzing laboratory
         reagent blanks as described in Section 9.3.1.  Subtracting blank values from  sample
         results is not permitted.

   4.3   Matrix interferences may be caused by contaminants that are co-extracted from the
         sample.  The extent of matrix interferences will vary considerably from source to

                                         540-6

-------
         source, depending upon the nature of the water. Humic and/or fulvic material can be
         co-extracted during SPE and high levels can cause enhancement and/or suppression in
         the electrospray ionization source or low recoveries on the SPE sorbent.3"4 Total
         organic carbon (TOC) is a good indicator of humic content of the sample.

   4.4   Relatively large quantities of the preservatives (Sect. 8.1.2) are added to sample
         bottles.  The potential exists for trace-level organic contaminants in these reagents.
         Interferences from these sources should be monitored by analysis of laboratory reagent
         blanks (Sect. 9.3.1), particularly when new lots of reagents are acquired.

   4.5   SPE cartridges can be a source of interferences. Analysis of field and laboratory
         reagent blanks can provide important information regarding the presence or absence of
         such interferences.  Brands and lots of SPE devices should be tested to ensure that
         contamination does not preclude analyte identification and quantitation.

5.  SAFETY

   5.1   The toxicity or carcinogenicity of each reagent used in this method has not been
         precisely defined. Each chemical should be treated as a potential health hazard, and
         exposure to these chemicals should be minimized. Each laboratory is responsible for
         maintaining an awareness of OSHA regulations regarding safe handling of chemicals
         used in this method.  A reference file of MSDSs should be made available to all
         personnel involved in the chemical analysis.  Additional references to laboratory
         safety are available.5"7

   5.2   Pure standard materials and stock standard solutions of these method analytes should
         be handled with suitable protection to skin and eyes, and care should be taken not to
         breathe the vapors or ingest the materials.

6.  EQUIPMENT AND SUPPLIES  (Brand names and/or catalog numbers are included for
   illustration only, and do not imply endorsement of the product.)

   6.1   SAMPLE CONTAINERS - 250-mL amber glass bottles fitted with
         polytetrafluoroethylene (PTFE)-lined screw caps.

   6.2   CENTRIFUGE TUBES  - 15-mL conical glass centrifuge tubes (Corning #8082-15) or
         other glassware suitable for collection of the eluent from the solid phase after
         extraction.

   6.3   AUTOSAMPLER VIALS -  Amber glass 2.0-mL autosampler vials (National
         Scientific #C4000-2W or equivalent) with caps containing PTFE-faced septa (National
         Scientific #C4000-53 or equivalent).

   6.4   MICRO SYRINGES - Suggested sizes include 5, 10, 25, 50, 100, 250,  500 and
         1000-jiL syringes.
                                          540-7

-------
6.5    ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g.

6.6    SOLID PHASE EXTRACTION (SPE) APPARATUS FOR USING CARTRIDGES

   6.6.1   SPE CARTRIDGES

      6.6.1.1  Waters Oasis HLB, 150 mg, 6cc (Waters # 186003365) - divinylbenzene N-
              vinylpyrrolidone copolymer.

      6.6.1.2  J.T. Baker Speedisk Column H2O-Philic DVB, 200 mg, 6cc (Baker
              # 8108-09) - modified divinylbenzene polymer.

   6.6.2   VACUUM EXTRACTION MANIFOLD

      6.6.2.1  Manual Extraction - A manual vacuum manifold with Visiprep™ large
              volume sampler (Supelco #57030 and #57275 or equivalent) for cartridge
              extractions.

      6.6.2.2  Automated Extraction - An automatic/robotic sample preparation system,
              designed for use with SPE cartridges, may be used if all QC requirements
              discussed in Section 9 are met. All sorbent washing, conditioning, sample
              loading, rinsing, drying and elution steps must be performed as closely as
              possible to the manual procedure.  Solvents used for washing, conditioning,
              and sample elution must be the same as those used in the manual procedure:
              however, the amount used may be increased as necessary to achieve the
              required data quality.  Solvent amounts may not be decreased.  Sorbent
              drying times prior to elution may be modified to achieve the required data
              quality. Caution should be exercised when increasing solvent volumes.
              Increased extract volume will likely necessitate the need for extended
              evaporation times which may compromise data quality. Caution should also
              be exercised when modifying sorbent drying times. Excessive drying may
              cause losses due to analyte volatility, and excessive contact with room air
              may oxidize some method analytes.

   6.6.3   SAMPLE DELIVERY SYSTEM - Use of a transfer tube system (Supelco
          "Visiprep," #57275 or equivalent), which transfers the sample directly from the
          sample container to the SPE cartridge is recommended.

6.7    EXTRACT CONCENTRATION SYSTEM - Extracts are concentrated by
      evaporation with nitrogen using a water bath set no higher than 40 °C (Meyer N-Evap,
      Model 111, Organomation Associates, Inc. or equivalent).

6.8    LABORATORY OR ASPIRATOR VACUUM SYSTEM - Sufficient capacity to
      maintain a vacuum of approximately 10 to 15 inches of mercury for extracting
      cartridges.
                                     540-8

-------
   6.9   LIQUID CHROMATOGRAPHY (LC)/TANDEM MASS SPECTROMETER
         (MS/MS) WITH DATA SYSTEM

       6.9.1  LC SYSTEM - Instrument capable of reproducibly injecting up to 10-|aL aliquots,
             and performing binary linear gradients at a constant flow rate near the flow rate
             used for development of this method (0.3 mL/min). Usage of a cooled
             autosampler compartment and a column heater is optional. Method performance
             data were collected using an autosampler thermostated to 4 °C during analyses.

       6.9.2  TANDEM MASS SPECTROMETER - The mass  spectrometer must be capable
             of positive ion  electrospray ionization (ESI) near the suggested LC flow rate of
             0.3 mL/min.  The system must be capable of performing MS/MS to produce
             unique product ions (Sect. 3.18) for method analytes within specified retention
             time segments. A minimum of 10 scans across the chromatographic peak is
             required to ensure adequate precision. Data demonstrated in Section 17 were
             collected using a triple quadrupole mass spectrometer.

       6.9.3  DATA SYSTEM - An interfaced data system is required to acquire, store, reduce,
             and output mass spectral data. The computer software should have the capability
             of processing stored LC/MS/MS data by recognizing an LC peak within any given
             retention time window. The software must allow integration of the ion abundance
             of any specific ion within specified time or scan number limits. The software
             must be able to calculate relative response factors,  construct linear regressions or
             quadratic calibration curves, and calculate analyte  concentrations.

       6.9.4  ANALYTICAL COLUMN - An LC Cig column (2.1 x 100 mm) packed with
             5 jam Cig solid phase particles (Restek Ultra Aqueous #9178512) was used.  Any
             equivalent column that provides adequate resolution, peak shape, capacity,
             accuracy, and precision (Sect. 9) may be used.

7.  REAGENTS AND STANDARDS

   7.1   GASES, REAGENTS, AND SOLVENTS - Reagent grade or better chemicals should
         be used. Unless otherwise indicated, it is intended that all reagents shall conform to
         the specifications of the Committee on Analytical Reagents of the American Chemical
         Society, where such specifications are available.  Other grades may be used, provided
         it is first determined that the reagent is  of sufficiently high purity to permit its use
         without lessening the quality of the determination.

       7.1.1  REAGENT WATER - Purified water which does not contain any measurable
             quantities of any method analytes or interfering compounds greater than 1/3 the
             MRL for each method analyte of interest.

       7.1.2  METHANOL (CH3OH, CAS#: 67-56-1) - High purity, demonstrated to be free of
             analytes and interferences (Fisher Optima LC/MS grade or equivalent).
                                        540-9

-------
7.1.3  ACETONITRILE (CH3CN, CAS#: 75-05-8) - High purity, demonstrated to be
      free of analytes and interferences (Tedia Absolv grade or equivalent).

7.1.4  ACETONE [(CH3)2CO, CAS#: 67-64-1] - High purity, demonstrated to be free of
      analytes and interferences (Tedia Absolv grade or equivalent).

7.1.5  FORMIC ACID (CH2O2; CAS# 64-18-6) - High purity, demonstrated to be free
      of analytes and interferences (Sigma-Aldrich ACS grade or equivalent).

7.1.6  AMMONIUM FORMATE (CH5O2N, CAS# 540-69-2) - High purity,
      demonstrated to be free of analytes and interferences (Fluka LC/MS grade or
      equivalent).

7.1.7  10 mM FORMATE BUFFER - To prepare 1 L, add 0.63 g ammonium formate
      and 0.5 mL formic acid to 1 L of reagent water. This solution is prone to
      volatility losses and should be replaced at least every 48 hours.

7.1.8  SAMPLE PRESERVATION REAGENTS - The following preservatives are
      solids at room temperature and may be added to the sample bottle before
      shipment to the field.

   7.1.8.1  TRIZMA PRESET CRYSTALS, pH 7.0  (Sigma-Aldrich #T-7193 or
           equivalent) - Reagent grade.  A premixed blend of Tris [Tris(hydroxy-
           methyl)aminomethane] and Tris HCL [Tris(hydroxymethyl)aminomethane
           hydrochloride]. Alternatively, a mix of the two components with a weight
           ratio of 15.5/1 Tris HCL/Tris may be used. These blends are targeted to
           produce a pH near 7.0 at 25 °C in reagent water. Trizma functions as a
           buffer (Sect. 8.1.2).

   7.1.8.2  L-ASCORBIC ACID (CAS# 50-81-7) -  Ascorbic acid reduces free chlorine
           at the time of sample collection (Sigma-Aldrich #255564 or equivalent).8

   7.1.8.3  2-CHLOROACETAMIDE (CAS# 79-07-2) - Inhibits microbial growth and
           analyte degradation (Sigma-Aldrich #C0267 or equivalent).8

7.1.9  NITROGEN - Aids in aerosol generation of the ESI liquid spray and is used as
      collision gas in some MS/MS instruments. Nitrogen used should meet or exceed
      instrument manufacturer's specifications.

7.1.10 ARGON - Used as collision gas during MS/MS experiments. Argon should meet
      or exceed instrument manufacturer's specifications. Nitrogen gas may be  used as
      collision gas provided sufficient sensitivity (product ion formation) is achieved.

  STANDARD SOLUTIONS - When the purity of a compound is assayed to be 96% or
  greater, the weight can be used without correction to calculate concentration of the
  stock standard.  The suggested concentrations are a description of concentrations used

                                 540-10

-------
  during method development, and may be modified to conform to instrument
  sensitivity. Standards for sample fortification generally should be prepared in the
  smallest volume that can be accurately measured to minimize addition of excess
  organic solvent to aqueous  samples.  Even though stability times for standard
  solutions are suggested in the following sections, laboratories should use standard
  QC practices to determine when their standards need to be replaced.

7.2.1  INTERNAL (IS) STOCK STANDARD SOLUTIONS - This method uses three
      IS compounds:  carbofuran-13Ce, bensulide-t/i4 and phorate-^io. These IS
      compounds were carefully chosen during method development because they
      encompass some of the functional groups of method analytes.  Although alternate
      IS compounds may be used provided they are isotopically labeled compounds
      with similar functional groups as method analytes, the analyst must have
      documented reasons for using alternate IS compounds.  Alternate  IS compounds
      must meet the QC requirements in  Section 9.3.4.

   7.2.1.1  IS STOCK STANDARD SOLUTIONS - These IS stocks can be obtained
           as individual certified stock standard solutions or neat materials.  During
           development of this method,  commercially obtained 100 |ig/mL stock
           standard solutions of carbofuran-13Ce in 1,4-dioxane (Cambridge Isotopes #
           CLM-1911-1.2) and phorate-dio in acetone (Crescent Chemical #
           XA16080100AC) were used. Bensulide-t/i4 was prepared from neat
           material (Cambridge Isotopes #DLM-7152) at 1000 |ig/mL in acetonitrile.
           IS stock standard solutions were stable for at least six months when stored at
           -5 °C or less in amber glass screw cap vials.

   7.2.1.2  INTERNAL STANDARD PRIMARY DILUTION STANDARD (IS PDS)
           (0.40 ng/|iL) - Prepare, or purchase commercially, the IS PDS at a
           suggested concentration of 0.40 ng/jiL using the three isotopically labeled
           chemicals in the table below. If prepared from individual stock standard
           solutions (Sect. 7.2.1.1), the table below can be used as a guideline for
           preparing the IS PDS although concentrations may need to be adjusted for
           instrument sensitivity.  The IS PDS used in these studies was prepared in
           acetonitrile. The IS PDS has been shown to be stable for at  least six months
           when stored at 6  °C or less in amber glass screw cap vials.  Ten jiL of this
           0.40 ng/|iL IS PDS was used to fortify the final 1-mL extracts (Sect. 11.4).
           This will yield a concentration of 4 pg/|iL of each IS in 1-mL extracts.


IS
carbofuran-13Ce
bensulide-t/i4
phorate-t/io
Cone, of
IS Stock
(ug/mL)
100
1000
100
Vol. Of
IS Stock
(uL)
20
2.0
20
Final Vol.
of IS PDS
(mL)
5.0
5.0
5.0
Final Cone.
of IS PDS
(ng/uL)
0.40
0.40
0.40
                                  540-11

-------
7.2.2   SURROGATE (SUR) ANALYTE STANDARD SOLUTIONS - The two SUR(s)
       for this method are methomyl-13C2,15N and tebuconazole-de-  These isotopically
       labeled SUR standards were carefully chosen during method development
       because they encompass some of the functional groups, as well as the water
       solubility range of method analytes.  Although alternate SUR standards may be
       used provided they are isotopically labeled compounds with similar functional
       groups as method analytes, the analyst must have documented reasons for using
       alternate SUR standards.  Alternate SUR standards chosen must still span the
       water solubility range of method analytes.  In addition, alternate SUR standards
       must meet the QC requirements in Section 9.3.5.

   7.2.2.1  SUR STOCK STANDARD SOLUTIONS - These SUR stocks can be
           obtained as individual certified stock standard solutions. During
           development of this method, commercially obtained 100 |ig/mL stock
           standard solutions of methomyl-13C2,15Nin methanol (Cambridge Isotopes
           #CNLM-7148-1.2) and tebuconazole-^4 in acetone (Dr. Ehrenstorfer GmbH
           #XA17178710AC) were used. SUR stock standard solutions were stable for
           at least one year when stored at -5  °C or less in amber glass screw cap vials.

   7.2.2.2  SURROGATE PRIMARY DILUTION STANDARD (SUR PDS)
           (0.40 ng/|iL) - Prepare, or purchase commercially, the SUR PDS  at a
           suggested concentration of 0.40 ng/|iL. If prepared from individual stock
           standard solutions (Sect. 7.2.2.1), the table below can be used as a guideline
           for preparing the SUR PDS. The SUR PDS used in these studies was
           prepared in acetonitrile.  This solution is used to fortify all QC and Field
           Samples. The PDS has been shown to be stable for at least six months when
           stored at 6 °C or less. Use 10 |iL of this 0.40 ng/jiL SUR PDS to fortify the
           250 mL aqueous QC and Field Samples prior to extraction (Sect.  11.2.2).
           This will yield  a concentration of 16 ng/L of each SUR in aqueous QC and
           Field Samples.


SUR
methomyl-13C2, 15N
tebuconazole-de
Cone. Of
SUR Stock
(ug/mL)
100
100
Vol. of
SUR Stock
GiL)
20
20
Final Vol. of
SUR PDS
(mL)
5.0
5.0
Final Cone.
of SUR PDS
(ng/uL)
0.40
0.40
 7.2.3 ANALYTE STANDARD SOLUTIONS - Analyte standards may be purchased
      commercially as ampulized solutions or prepared from neat materials (see Table 3
      for analyte sources used during method development).

   7.2.3.1  ANALYTE STOCK STANDARD SOLUTION (1000 |ig/mL) - If
           preparing from neat material, accurately weigh approximately 5 mg of pure
           material to the nearest 0.1 mg and dilute to 5 mL with acetonitrile or
                                 540-12

-------
            methanol for a final concentration of 1000 ug/mL. Repeat for each method
            analyte prepared from neat material. Alternatively, purchase commercially
            available individual stock standards of the analytes, preferably in methanol or
            acetonitrile, if available. For development of this method, commercially
            available stock standards of 1000 ug/mL were purchased except for phorate
            sulfoxide, tebufenozide, disulfoton sulfoxide and chlorpyrifos oxon which
            were purchased as neat materials. These stock standards were stable for at
            least six months when stored at -5 °C or less in amber glass screw cap vials.

   7.2.3.2   ANALYTE PRIMARY DILUTION STANDARD (PDS) SOLUTION
            (0.16-0.40  ng/|aL) - The analyte PDS contains all, or a portion, of method
            analytes at various concentrations in acetonitrile.  ESI and MS/MS response
            varies by compound; therefore, a mix of concentrations may be needed in
            the analyte PDS.  During method development, analyte PDS solutions were
            prepared such that approximately the same instrument response was
            obtained for all analytes.  The analyte PDS was prepared in acetonitrile at
            concentrations of 0.4 ng/|aL, except for fenamiphos, tebufenozide and
            tebuconazole at 0.16 ng/|aL each. The analyte PDS is prepared by dilution
            of the combined Analyte Stock Standard Solutions (Sect.7.2.3.1) and is used
            to prepare CAL standards, and fortify LFBs, LFSMs, LFSMDs and FDs
            with the method analytes.  The analyte PDS has been shown to be stable for
            six months when  stored at 6 °C or less in amber glass screw cap vials.

7.2.4   CALIBRATION STANDARDS (CAL) - Prepare a series of at least five
       concentrations of calibration solutions in methanol (see note below), from
       dilutions of the analyte PDS (Sect 7.2.3.2).  The suggested concentrations in this
       paragraph are a description of the concentrations used during method
       development, and may be modified to conform with instrument  sensitivity.
       Concentrations ranging from 0.16-6.4 ug/L are suggested for fenamiphos,
       tebufenozide and tebuconazole, and 0.40-16 ug/L are suggested for the remaining
       analytes.  Larger concentration ranges will require more calibration points.  The
       IS and SUR are added to CAL standards at a constant concentration. During
       method development, the concentrations of the SUR(s) were 0.40 ng/|aL in the
       standard (16 ng/L in the aqueous sample) and the IS(s) concentrations were
       4 pg/|-iL.  The lowest concentration  CAL standard must be at or below the MRL,
       which may depend on system sensitivity.  CAL standards may also be used as
       CCCs (Sect. 9.3.2). During method development, CAL standards were shown to
       be stable for two weeks when stored at 6 °C or less. Longer storage times are
       acceptable provided appropriate QC measures are  documented demonstrating the
       CAL stability.

       Note: Acetonitrile was not used as the solvent for calibration standards
             because injection of calibration standards prepared in acetonitrile
             distorted peak shapes for early eluting analytes, such  as methomyl and
             3-hydroxycarbofuran.
                                  540-13

-------
8.  SAMPLE COLLECTION, PRESERVATION, AND STORAGE

   8.1   SAMPLE BOTTLE PREPARATION

       8.1.1  Samples must be collected in 250-mL amber glass bottles fitted with teflon-lined
             screw caps.

       8.1.2  Preservation reagents, listed in the table below, are added to each sample bottle as
             a solid prior to shipment to the field (or prior to sample collection).
Compound
Trizma
2-Chloroacetamide
Ascorbic acid
Amount
7.75 g/L
2g/L
100 mg/L
Purpose
buffering reagent
antimicrobial
dechlorinating agent
   8.2   SAMPLE COLLECTION

       8.2.1  Open the tap and allow the system to flush until the water temperature has
             stabilized (approximately 3 to 5 min).  Collect samples from the flowing system.

       8.2.2  Fill sample bottles, taking care not to flush out the sample preservation reagents.
             Samples do not need to be collected headspace free.

       8.2.3  After collecting the sample, cap the bottle and agitate by hand until preservative is
             dissolved. Note that 2-chloroacetamide is slow to dissolve especially in cold
             water.  Keep the sample sealed from time of collection until extraction.

   8.3   SAMPLE SHIPMENT AND  STORAGE - Samples must be chilled during shipment
         and must not exceed 10 °C during the first 48 hours after collection. Sample
         temperature must be confirmed to be at or below 10 °C when samples are received at
         the laboratory. Samples stored in the lab must be held at or below 6 °C until
         extraction, but should not be frozen.

         NOTE: Samples that are significantly above 10°  C, at the time of collection, may need
                to be iced or refrigerated for a period of time, in order to chill them prior to
                shipping. This will allow them to be shipped with sufficient ice to meet the
                above requirements.

   8.4   SAMPLE AND EXTRACT HOLDING TIMES - Water samples should be extracted
         as soon as possible after collection but must be extracted within 28 days of collection,
         except for chlorpyrifos oxon which must be extracted within 7 days. Extracts must be
         stored at < 6 °C and analyzed within 28 days after extraction.
                                        540-14

-------
9.  QUALITY CONTROL

   9.1   QC requirements include the Initial Demonstration of Capability (IDC) and ongoing
         QC requirements that must be met when preparing and analyzing Field Samples. This
         section describes QC parameters, their required frequencies, and performance criteria
         that must be met in order to meet EPA quality objectives.  QC criteria discussed in the
         following sections are summarized in Tables 14 and 15. These QC requirements are
         considered the minimum acceptable QC criteria.  Laboratories are encouraged to
         institute additional QC practices to meet their specific needs.

       9.1.1  METHOD MODIFICATIONS - The analyst is permitted to modify LC columns,
             LC conditions, evaporation techniques, internal standards or surrogate standards,
             and MS and MS/MS conditions.  Each time such method modifications are made,
             the analyst must repeat the procedures of the IDC. Modifications to LC
             conditions  should still minimize co-elution of method analytes to reduce the
             probability of suppression/enhancement effects.

   9.2   INITIAL DEMONSTRATION OF CAPABILITY (IDC)  - The IDC must be
         successfully performed prior to analyzing any Field Samples. Prior to conducting the
         IDC, the analyst must first generate an acceptable Initial Calibration following the
         procedure outlined in Section 10.2.

       9.2.1  INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND - Any time
             a new lot of SPE cartridges, solvents, centrifuge tubes, disposable pipets, and
             autosampler vials are used,  it must be demonstrated that an LRB is reasonably
             free of contamination and that criteria in Section 9.3.1 are met.  If an automated
             extraction system is used, an LRB should be extracted on each port to ensure that
             all valves and tubing are free from potential contamination.

       9.2.2  INITIAL DEMONSTRATION OF PRECISION (IDP) - Prepare, extract, and
             analyze four to seven replicate LFBs fortified near the midrange of the initial
             calibration curve according to the procedure described in Section 11.  Sample
             preservatives as described in Section 8.1.2 must be added to these samples.  The
             relative standard deviation (RSD) of the results  of replicate analyses must be less
             than 20%.

       9.2.3  INITIAL DEMONSTRATION OF ACCURACY (IDA) - Using the same set of
             replicate data generated for Section 9.2.2, calculate average recovery. The
             average recovery of replicate values must be within ± 30%  of the true value.

       9.2.4  MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish a target
             concentration for the MRL based on the intended use of the method. The MRL
             may be established by a laboratory for their specific purpose or may be set by a
             regulatory agency.  Establish an Initial Calibration following the procedure
             outlined in  Section 10.2.  The lowest CAL standard used to establish Initial
             Calibration (as well as the low-level CCC, Section 10.3) must be at or below the

                                        540-15

-------
       concentration of the MRL. Establishing the MRL concentration too low may
       cause repeated failure of ongoing QC requirements.  Confirm the MRL following
       the procedure outlined below.

   9.2.4.1  Fortify, extract, and analyze seven replicate LFBs at the proposed MRL
           concentration. These LFBs must contain all method preservatives described
           in Section 8.1.2. Calculate the mean measured concentration (Mean) and
           standard deviation for these replicates. Determine the Half Range for the
           prediction interval of results (HRpiR) using the equation below
                                        HRPIR = 3.963s

            where
                   s      = standard deviation
                   3.963  = a constant value for seven replicates.1

   9.2.4.2   Confirm that the upper and lower limits for the Prediction Interval of Result
            (PIR = Mean j_ HRpiR) meet the upper and lower recovery limits as shown
            below

            The Upper PIR Limit must be < 150% recovery.

                            Mean + HRP1R
                        FortifiedConcentration

            The Lower PIR Limit must be > 50% recovery.
                             Mean-HRP1R
                         FortifiedConcentration

   9.2.4.3   The MRL is validated if both the Upper and Lower PIR Limits meet the
            criteria described above (Sect. 9.2.4.2).  If these criteria are not met, the
            MRL has been set too low and must be determined again at a higher
            concentration.

9.2.5   CALIBRATION CONFIRMATION - Analyze a QCS as described in Section
       9.3.8 to confirm the accuracy of the standards/calibration curve.

9.2.6   DETECTION LIMIT DETERMINATION (optional) - While DL determination
       is not a specific requirement of this method, it may be required by various
       regulatory bodies associated with compliance monitoring. It is the responsibility
       of the laboratory to determine ifDL determination is required based upon the
       intended use of the data.
                                  540-16

-------
          Replicate analyses for this procedure should be done over at least three days (i.e.,
          both the sample extraction and the LC/MS/MS analyses should be done over at
          least three days).  Prepare at least seven replicate LFBs at a concentration
          estimated to be near the DL. This concentration may be estimated by selecting a
          concentration at 2-5 times the noise level.  DLs in Table 5 were calculated from
          LFBs fortified at various concentrations as indicated in the table.  Appropriate
          fortification concentrations will be dependent upon the sensitivity of the
          LC/MS/MS system used.  All preservation reagents listed in Section 8.1.2 must
          also be added to these samples. Analyze the seven replicates through all steps of
          Section 1 1 .

          NOTE: If an MRL confirmation data set meets these requirements, a DL may be
                  calculated from the MRL confirmation data, and no additional analyses
                  are necessary.

          Calculate the DL using the following equation

                                         — J A, t
                                              /  i i     /-. /-\/-\\
                                              (w-l,l-a=0.99)

                       where
                              s = standard deviation of replicate analyses
                              t («-i, i-a=o.99) = Student's t value for the 99% confidence
                                           level with n-1 degrees of freedom
                              n = number of replicates.

          NOTE: Do not subtract blank values when performing DL calculations.

9.3    ONGOING QC REQUIREMENTS - This section summarizes ongoing QC criteria
      that must be followed when processing and analyzing Field Samples.

   9.3.1   LABORATORY REAGENT BLANK (LRB) - An LRB is required with each
          extraction batch (Sect. 3.6) to confirm that potential background contaminants are
          not interfering with identification or quantitation  of method analytes. If more
          than 20 Field Samples are included in a batch, analyze an LRB for every 20
          samples.  If the LRB produces a peak within the retention time window of any
          analyte that would prevent determination of that analyte, determine the source of
          contamination and eliminate the interference before processing samples.
          Background contamination must be reduced to an acceptable level before
          proceeding. Background from method analytes or other contaminants that inter-
          fere with the measurement of method analytes must be below 1/3  of the MRL.
          Blank contamination is estimated by extrapolation, if the concentration is below
          the lowest CAL standard. This extrapolation procedure is not allowed for sample
          results as it may not meet data quality objectives. If method analytes are detected
          in the LRB at concentrations equal to or greater than this level, then all data for
          the problem analyte(s) must be considered invalid for all samples  in the extraction
          batch.
                                      540-17

-------
9.3.2   CONTINUING CALIBRATION CHECK (CCC) - CCC standards are analyzed
       at the beginning of each analysis batch, after every 10 Field Samples, and at the
       end of the analysis batch. See Section 10.3 for concentration requirements and
       acceptance criteria.

9.3.3   LABORATORY FORTIFIED BLANK (LFB) - An LFB is required with each
       extraction batch (Sect. 3.6). The fortified concentration of the LFB must be
       rotated between low, medium, and high concentrations from batch to batch. The
       low concentration LFB must be as near as practical to, but no more than two
       times, the MRL. Similarly, the high concentration LFB should be near the high
       end of the calibration range established during the initial calibration (Sect.  10.2).
       Results of low-level LFB analyses must be 50-150% of the true value. Results of
       medium and high-level LFB analyses must be 70-130% of the true value. If LFB
       results do not meet these criteria for method analytes, then all data for the
       problem analyte(s) must be considered invalid for all samples in the extraction
       batch.

9.3.4   INTERNAL STANDARDS (IS) - The analyst must monitor peak areas  of the
       IS(s) in all injections during each analysis day.  Internal standard responses (as
       indicated by peak areas) for any chromatographic run must not deviate by more
       than ± 50%  from average areas measured during the initial calibration for the
       internal standards.  If IS areas in a chromatographic run do not meet these criteria,
       inject a second aliquot of that standard or extract.

   9.3.4.1   If the reinjected aliquot produces an acceptable IS response, report results
            for that aliquot.

   9.3.4.2   If the reinjected extract fails again,  the analyst should check the calibration
            by reanalyzing the most recently acceptable CAL standard.  If the CAL
            standard fails the criteria of Section 10.3, recalibration is in order per
            Section 10.2. If the CAL standard is acceptable, extraction of the sample
            may need to be repeated provided the sample is still within the holding time.
            Otherwise, report results obtained from  the reinjected extract, but annotate
            as suspect. Alternatively,  collect a  new sample and re-analyze.

9.3.5   SURROGATE RECOVERY - The SUR standard is fortified into all samples,
       CCCs, LRBs, LFBs, LFSMs, LFSMDs, and FD prior to extraction.  It is also
       added to CAL standards. The SUR is a means of assessing method performance
       from extraction to final chromatographic measurement.  Calculate the  recovery
       (%R) for the SUR using the following equation

                                      ( A\
                                 %R= — xlOO
                                      UJ
       where
             A   = calculated SUR concentration for the QC or Field Sample

                                  540-18

-------
             B   = fortified concentration of the SUR.

   9.3.5.1   SUR recovery must be in the range of 70-130%. When SUR recovery from
            a sample, blank, or CCC is less than 70% or greater than  130%, check 1)
            calculations to locate possible errors, 2) standard solutions for degradation,
            3) contamination, and 4) instrument performance.  Correct the problem and
            reanalyze the extract.

   9.3.5.2   If the extract reanalysis meets the SUR recovery criterion, report only data
            for the reanalyzed extract.

   9.3.5.3   If the extract reanalysis fails the 70-130% recovery criterion, the analyst
            should check the calibration by injecting the last CAL standard that passed.
            If the CAL standard fails the criteria of Section 10.3, recalibration is in
            order per Section 10.2. If the CAL standard is acceptable, extraction of the
            sample should be repeated provided the sample is still within the holding
            time.  If the re-extracted sample also fails the recovery criterion, report all
            data for that sample as suspect/SUR recovery to inform the data user that the
            results are suspect due to SUR recovery. Alternatively, collect a new
            sample and re-analyze.

9.3.6   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of an
       LFSM is required in each extraction batch and is used to determine that the
       sample matrix does not adversely affect method accuracy. Assessment of method
       precision is accomplished by analysis of a Field Duplicate (FD) (Sect. 9.3.7);
       however, infrequent occurrence of method analytes would hinder this assessment.
       If the occurrence of method analytes in samples is infrequent, or if historical
       trends are unavailable, a second LFSM, or LFSMD, must be prepared, extracted,
       and analyzed from a duplicate of the Field Sample.  Extraction batches that
       contain LFSMDs will not require extraction of a FD. If a variety of different
       sample matrices are analyzed regularly, for example, drinking water from ground
       water and surface water sources, method performance should be established for
       each.  Over time, LFSM data should be documented by the laboratory for all
       routine sample sources.

   9.3.6.1   Within each extraction batch (Sect. 3.6), a minimum of one Field Sample is
            fortified as an LFSM for every 20 Field Samples analyzed.  The LFSM is
            prepared by spiking a sample with an appropriate amount of the Analyte
            PDS (Sect. 7.2.3.2).  Select a spiking concentration that is greater than or
            equal to the matrix background concentration, if known.  Use historical data
            and rotate through low, mid and high concentrations when selecting a
            fortifying concentration.

   9.3.6.2   Calculate percent recovery (%R) for each analyte using the equation
                                  540-19

-------
                                         c
   where
                   A   = measured concentration in the fortified sample
                   B   = measured concentration in the unfortified sample
                   C   = fortification concentration.

   9.3.6.3    Analyte recoveries may exhibit matrix bias. For samples fortified at or
            above their native concentration, recoveries should range between 70-130%,
            except for low-level fortification near or at the MRL (within a factor of
            two-times the MRL concentration) where 50-150% recoveries are
            acceptable. If the accuracy of any analyte falls outside the designated range,
            and laboratory performance for that analyte is shown to be in control in
            CCCs, the recovery is judged to be matrix biased. The result for that analyte
            in the unfortified sample is labeled suspect/matrix to inform the data user
            that the results are suspect due to matrix effects.

9.3.7   FIELD DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
       DUPLICATE (FD or LFSMD) - Within each extraction batch (not to exceed 20
       Field Samples, Sect. 3.6), a minimum of one FD or LFSMD must be analyzed.
       Duplicates check the precision associated with sample collection, preservation,
       storage, and laboratory procedures.  If method analytes are not routinely observed
       in Field Samples, an LFSMD should be analyzed rather than an FD.

   9.3.7.1   Calculate relative percent difference (RPD) for duplicate measurements
            (FD1 and FD2) using the equation

                                FDI-FD2
                       RPD = -/	,— xlOO
                              (FDl + FD2)/2

   9.3.7.2   RPDs for FDs should be < 30%. Greater variability may be observed when
            the matrix is fortified at analyte concentrations at or near the MRL (within a
            factor of two times the MRL concentration).  At these concentrations, FDs
            should have RPDs that  are < 50%. If the RPD of any analyte falls outside
            the designated range, and laboratory performance for that analyte is shown
            to be in control in the CCC, the recovery is judged to be matrix biased. The
            result for that analyte in the unfortified sample is labeled suspect/matrix to
            inform the data user that the results are suspect due to matrix effects.

   9.3.7.3   If an LFSMD is analyzed instead of a FD, calculate the relative percent
            difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the
            equation

                             \LFSM-LFSMD\
                           ^	^
                            (LFSM+LFSMD} 12

                                  540-20

-------
          9.3.7.4  RPDs for duplicate LFSMs should be < 30% for samples fortified at or
                  above their native concentration. Greater variability may be observed when
                  the matrix is fortified at analyte concentrations at or near the MRL (within a
                  factor of two times the MRL concentration). LFSMs fortified at these
                  concentrations should have RPDs that are < 50% for samples fortified at or
                  above their native concentration.  If the RPD of any analyte falls outside the
                  designated range, and laboratory performance for that analyte is shown to be
                  in control in the CCC, the recovery is judged to be matrix biased. The result
                  for that analyte in the unfortified sample is labeled suspect/matrix to inform
                  the data user that the results are suspect due to matrix  effects.

       9.3.8  QUALITY CONTROL SAMPLES (QCS) - As part of the IDC (Sect. 9.2), each
             time a new Analyte PDS (Sect. 7.2.3.2) is prepared, and at least quarterly, analyze
             a QCS sample from  a source different from the source of the CAL standards.  If a
             second vendor is not available, then a different lot of the standard should be used.
             The QCS should be prepared and analyzed just like a CCC.  Fortify the QCS near
             the midpoint of the calibration range.  Acceptance criteria for the QCS are  identical
             to mid- and high-level CCCs; the calculated amount for each analyte must be
             ± 30% of the true value.  If measured analyte concentrations are not of acceptable
             accuracy, check the entire analytical procedure to locate and correct the problem.
             If the discrepancy is not resolved, one of the standard materials may be degraded
             or otherwise compromised and a third standard must be obtained.

10.  CALIBRATION AND STANDARDIZATION

    10.1  Demonstration and documentation  of acceptable initial calibration is required before
         any samples are analyzed.  The MS tune check and initial calibration must be  repeated
         each time a major instrument modification is made, or maintenance is performed.

    10.2  INITIAL CALIBRATION

       10.2.1 ESI-MS/MS TUNE

          10.2.1.1 Calibrate the mass scale  of the MS with the calibration compounds  and
                  procedures prescribed by the manufacturer.

          10.2.1.2 Optimize the [M+H]+ for each method analyte by infusing approximately
                  0.5-1.0 ng/mL of each analyte (prepared in the initial mobile phase
                  conditions) directly into the MS at the chosen LC mobile phase flow rate
                  (approximately 0.3 mL/min). This tune can be done on a mix of method
                  analytes. MS parameters (voltages, temperatures, gas flows, etc.) are varied
                  until optimal analyte responses are determined. Method analytes may have
                  different optima  requiring some compromise between the optima.  See
                  Table 2 for ESI-MS conditions used in method development.
                                         540-21

-------
   10.2.1.3  Optimize the product ion (Sect. 3.18) for each analyte by infusing
            approximately 0.5-1.0 ng/mL of each analyte (prepared in the initial mobile
            phase conditions) directly into the MS at the chosen LC mobile phase flow
            rate (approximately 0.3 mL/min).  This tune can be done on a mix of
            method analytes. MS/MS parameters (collision gas pressure, collision
            energy, etc.) are varied until optimal analyte responses are determined. See
            Table 4 for MS/MS conditions used in method development.

10.2.2 Establish LC operating parameters that optimize resolution and peak shape.
      Suggested LC conditions can be found in Table 1.  LC conditions listed in Table 1
      may not be optimum for all LC systems and may need to be optimized by the
      analyst.

10.2.3 Inject a mid-level CAL standard under LC/MS conditions to obtain retention
      times of each method analyte. Divide the chromatogram into retention time
      windows (segments) each of which contains one or more chromatographic peaks.
      During MS/MS analysis, fragment a small number of selected precursor ions
      ([M+H]+; Sect. 3.16) for the analytes in each window and choose the most
      abundant product ion. Product ions (also quantitation ions) chosen during method
      development are in Table 4, although these will be instrument dependent. For
      maximum sensitivity in subsequent MS/MS analyses, minimize the number of
      transitions that are simultaneously monitored within each segment.

10.2.4 Inject a mid-level CAL standard under optimized LC/MS/MS conditions  to ensure
      that each method analyte is observed in its MS/MS window and that there are at
      least 10 scans across the peak for optimum precision.

10.2.5 Prepare a set of at least five CAL standards as described in Section 7.2.4. The
      lowest concentration CAL standard must be at or below the MRL,  which  may
      depend on system sensitivity. It is recommended that at least four  of the CAL
      standards are at a concentration greater than or equal to the MRL.

10.2.6 The LC/MS/MS system is calibrated using the IS technique. Use the LC/MS/MS
      data system software to generate a linear regression or quadratic calibration curve
      for each of the analytes. Curves may be concentration weighted, if necessary.

10.2.7 CALIBRATION ACCEPTANCE CRITERIA - Validate the initial calibration by
      calculating the concentration of each analyte as an unknown against its regression
      equation. For calibration levels that are < MRL, the result for each analyte should
      be within ± 50% of the true value. All other calibration  points must calculate to be
      within ± 30% of their true value.  If these criteria cannot be met, the analyst will
      have difficulty meeting ongoing QC criteria. It is recommended that corrective
      action is taken to reanalyze the CAL standards, restrict the range of calibration, or
      select an alternate method of calibration.
                                  540-22

-------
          CAUTION: When acquiring MS/MS data, LC operating conditions must be
                      carefully reproduced for each analysis to provide reproducible
                      retention times. If this is not done, the correct ions will not be
                      monitored at appropriate times. As a precautionary measure,
                      chromatographic peaks in each window must not elute too close to
                      the edge of the segment time window.

10.3   CONTINUING CALIBRATION CHECK (CCC) - Minimum daily calibration
      verification is as follows.  Analyze a CCC to verify the initial calibration at the
      beginning of each analysis batch, after every tenth Field Sample, and at the end of
      each analysis batch.  LRBs, CCCs, LFBs, LFSMs, FDs and LFSMDs are not counted
      as samples. The beginning CCC of each analysis batch must be at or below the MRL
      in order to verify instrument sensitivity prior to any analyses. If standards have been
      prepared such that all low CAL points are not in the same CAL solution, it may be
      necessary to analyze two CAL standards to meet this  requirement.  Alternatively,
      analyte concentrations in the analyte PDS may be customized to meet this criteria.
      Subsequent CCCs should alternate between a medium and high concentration CAL
      standard.

   10.3.1  Inject an aliquot of the appropriate concentration CAL standard and analyze with
          the same conditions used during the initial calibration.

   10.3.2  Determine that the absolute areas of the quantitation ions of the IS(s) are within
          50-150% of the average areas measured during initial calibration. If any of the IS
          areas has changed by more than these amounts, adjustments must be made to
          restore system sensitivity. These adjustments may include cleaning of the MS ion
          source, or other maintenance as indicated in Section 10.3.4. Major instrument
          maintenance requires recalibration (Sect 10.2) and verification of sensitivity by
          analyzing a CCC at or below the MRL (Sect 10.3).  Control charts are useful aids
          in documenting system sensitivity changes.

   10.3.3  Calculate the concentration of each analyte and SUR in the CCC. The calculated
          amount for the  SUR must be within ± 30% of the true value.  Each analyte
          fortified at a level < MRL must calculate to be within ± 50% of the true value.
          The calculated  concentration of method analytes in CCCs fortified at all other
          levels must be within ± 30%.  If these conditions do not exist, then all data for the
          problem analyte must be considered invalid, and remedial action should be taken
          (Sect. 10.3.4) which may require recalibration. Any Field or QC Samples that
          have been analyzed since the last acceptable calibration verification should be
          reanalyzed after adequate calibration has been restored, with the following
          exception. If the CCC fails because the calculated concentration is greater
          than 130% (150% for the low-level CCC) for a particular method analyte,
          and Field Sample extracts show no detection for that method analyte, non-
          detects may be reported without re-analysis.
                                     540-23

-------
      10.3.4 REMEDIAL ACTION - Failure to meet CCC QC performance criteria may
            require remedial action. Major maintenance, such as cleaning the electrospray
            probe, atmospheric pressure ionization source, mass analyzer, replacing the LC
            column, etc., requires recalibration (Sect 10.2) and verification of sensitivity by
            analyzing a CCC at or below the MRL (Sect 10.3).

11.  PROCEDURE

   11.1  This procedure may be performed manually or in an automated mode using a robotic
        or automatic sample preparation device.  Data presented in Tables 5-13 demonstrate
        data collected by manual extraction. If an automated system is used to prepare
        samples, follow the manufacturer's operating instructions, but all extraction and
        elution steps must be the same as  in the manual procedure. Extraction and/or elution
        steps may not be changed or omitted to accommodate the use of an automated system.
        If an automated system is used, LRBs should be rotated among the ports to ensure that
        all valves and tubing meet LRB requirements (Sect. 9.3.1).

        NOTE:  SPE cartridges described in this section are designed as single use items  and
                 should be discarded after use. They may not be refurbished for reuse in
                 subsequent analyses.

   11.2  SAMPLE PREPARATION

      11.2.1 Samples are preserved, collected and stored as presented in Section 8. All Field
            and QC Samples, including the LRB, and LFB, must contain the preservatives
            listed in Section 8.1.2. Before extraction, verify that the sample pH is 7 ± 0.5. If
            the sample pH does not meet this requirement, discard the sample. If the  sample
            pH is acceptable, proceed with the analysis. Before extraction, mark the level of
            the sample on the outside of the sample bottle for later sample volume
            determination (Sect. 11.5). If using weight to determine volume, weigh the bottle
            with collected sample before extraction.

      11.2.2 Add an aliquot of the SUR PDS (Sect. 7.2.2.2) to each sample to be extracted, cap
            and invert to mix. During method development, a 10-|aL aliquot of the
            0.40 ng/nL SUR PDS  (Sect. 7.2.2.2) was added to 250 mL for a final
            concentration  of 16 ng/L in the aqueous sample.

      11.2.3 In addition to  SUR(s) and preservatives, if the sample is an LFB, FD, LFSM,  or
            LFSMD, add the necessary amount of analyte PDS (Sect. 7.2.3.2). Cap and invert
            each sample to mix.

    11.3 CARTRIDGE SPE PROCEDURE

      11.3.1 CARTRIDGE CLEAN-UP AND CONDITIONING - DO NOT allow cartridge
            packing material to go dry during any of the conditioning steps.  Rinse each
            cartridge with 5 mL of methanol.  Next, rinse each cartridge with 10 mL of

                                        540-24

-------
          reagent water, without allowing the water to drop below the top edge of the
          packing. If the cartridge goes dry during the conditioning phase, the conditioning
          must be started over. Add 4-5 mL of reagent water to each cartridge, attach
          sample transfer tubes (Sect. 6.6.3), turn on the vacuum, and begin adding sample
          to the cartridge.

   11.3.2  SAMPLE EXTRACTON - Adjust the vacuum so that the approximate flow rate
          is 10-15 mL/min. Do not allow the cartridge to go dry before all the sample has
          passed through.

   11.3.3  CARTRIDGE RINSE - After the entire sample has passed through the cartridge,
          rinse the cartridge with 5 mL of reagent water.  Draw air or nitrogen through the
          cartridge for 5 min at high vacuum (10-15 in. Hg).

   11.3.4  CARTRIDGE ELUTION - Turn off and release the vacuum. Lift the  extraction
          manifold top and insert a rack with collection tubes into the extraction tank to
          collect the extracts as they are eluted from the cartridges.  Elute analytes from the
          cartridge by pulling 5 mL of methanol through the cartridge. Use a low vacuum
          such that the solvent exits the cartridge in a dropwise fashion.

11.4  EXTRACT  CONCENTRATION - Concentrate the extract to less than 1 mL (but not
     less than 0.5 mL) under a gentle stream of nitrogen in a heated water bath (40 °C).
     Add the appropriate amount of methanol and IS PDS (Sect. 7.2.1.2) to the collection
     vial to bring the volume to 1 mL and vortex. (10 |aL of the 0.40 ng/|jL IS PDS for
     extract concentrations of 4 pg/|iL were used during method development).  Transfer a
     small aliquot to an autosampler vial.

11.5  SAMPLE VOLUME DETERMINATION - If the level of the sample was marked on
     the sample bottle, use a graduated cylinder to measure the volume of water required to
     fill the original sample bottle to the mark made prior to extraction.  Determine to the
     nearest 10 mL. If using weight to determine volume, weigh the empty bottle to the
     nearest 10 g and determine the  sample weight by subtraction of the empty bottle
     weight from the original sample weight (Sect. 11.2.1). Assume a sample density of
     1.0 g/mL. In either case, the sample volume will be used in the final calculations of
     the analyte concentration (Sect. 12.2).

11.6  EXTRACT ANALYSIS

   11.6.1  Establish operating conditions equivalent to those summarized in Tables 1-4 of
          Section 17.  Instrument conditions and columns should be optimized prior to
          initiation of the IDC.

   11.6.2  Establish an appropriate retention time window for each analyte. This should be
          based on measurements of actual  retention time variation for each method analyte
          in CAL standard solutions analyzed on the LC over the course of time.  A value
          of plus  or minus three times the standard deviation of the retention time obtained

                                     540-25

-------
              for each method analyte while establishing the initial calibration and completing
              the IDC can be used to calculate a suggested window size. However, the
              experience of the analyst should weigh heavily on the determination of the
              appropriate retention window size.

       11.6.3  Establish a valid initial calibration following the procedures outlined in Sect. 10.2
              or confirm that the calibration is still valid by running a CCC as described in Sect.
              10.3. If establishing an initial calibration, complete the IDC as described in
              Section 9.2.

       11.6.4  Begin analyzing Field Samples, including QC samples, at their appropriate
              frequency by injecting the same  size aliquots (10 |aL was used in method
              development), under the same conditions used to analyze the CAL standards.

       11.6.5  At the conclusion of data acquisition, use the same software that was used in the
              calibration procedure to identify peaks of interest in predetermined retention time
              windows. Use the data system software to examine the ion abundances of the
              peaks in the chromatogram. Identify an analyte by comparison of its retention
              time with that of the corresponding method analyte peak in a reference standard.
              Comparison of MS/MS mass spectra is not particularly useful given the limited
              ±0.5 dalton mass range around a single product ion for each method analyte.

       11.6.6  The analyst must not extrapolate beyond the established calibration range.  If an
              analyte peak area exceeds the range of the initial calibration curve, the extract
              may be diluted with methanol  and the appropriate amount of IS added to match
              the original concentration.  Re-inject the diluted extract. Incorporate the dilution
              factor into the final concentration calculations. Acceptable SUR performance
              (Sect. 9.3.5.1) should be determined from the undiluted sample extract. The
              resulting data should be documented as a dilution and MRLs should be adjusted
              accordingly.

12. DATA ANALYSIS AND CALCULATION

   12.1.   Complete chromatographic resolution is not necessary for accurate and precise
          measurements of analyte concentrations using MS/MS.  In validating this method,
          concentrations were calculated by measuring the product ions listed in Table 4.  Other
          ions may be selected at the discretion of the analyst.

   12.2.   Calculate analyte and SUR concentrations using the multipoint calibration established
          in Section 10.2. Do not use daily calibration verification data to quantitate analytes in
          samples.  Adjust final analyte concentrations to reflect the actual sample volume
          determined in Section 11.5.

   12.3  Prior to reporting data, the chromatogram should be reviewed for any incorrect peak
          identification or poor integration.
                                          540-26

-------
   12.4  Calculations must utilize all available digits of precision, but final reported
         concentrations should be rounded to an appropriate number of significant figures (one
         digit of uncertainty), typically two, and not more than three significant figures.

         NOTE: Some data in Section 17 of this method are reported with more than two
                 significant figures.  This is done to better illustrate method performance.

13. METHOD PERFORMANCE

   13.1  PRECISION, ACCURACY, AND MINIMUM REPORTING LEVELS - Tables for
         these data are presented in Section 17. LCMRLs and DLs for each method analyte are
         presented in Table 5.  Precision and accuracy are presented for three water matrices:
         reagent water (Tables 6  and 7); chlorinated (finished) ground water (Tables 8 and 9);
         chlorinated (finished) surface water (Tables 10 and 11).

   13.2  SAMPLE STORAGE STABILITY STUDIES - An analyte storage stability study was
         conducted by fortifying  the analytes into chlorinated surface water samples that were
         collected, preserved, and stored as described in Section 8. Precision and mean
         recovery (n=4) of analyses, conducted on Days 0, 7, 14, 23 and 28 are presented in
         Table 12.

   13.3  EXTRACT STORAGE  STABILITY STUDIES - Extract storage stability studies
         were conducted on extracts obtained from a chlorinated surface water fortified with
         method analytes. Precision and mean recovery (n=4) of injections conducted on Days
         0, 7, 14, 23, and 28 are reported in Table 13.

   13.4  SECOND LABORATORY DEMONSTRATION - Performance of this method was
         demonstrated by multiple laboratories, with results similar to those reported in Section
         17.  The authors wish to acknowledge the assistance of the analysts and laboratories
         for their participation in the multi-laboratory verification studies: a) Dr. Andrew
         Eaton and Mr. Ali Haghani of Eurofms Eaton Analytical (EEA), Monrovia, CA, b) Dr.
         Yongtao Li and Mr. Joshua S. Whitaker of UL LLC, South Bend, IN, and c) Dr.
         Damon Carl of Heritage Environmental Services, Indianapolis, IN.

14.  POLLUTION PREVENTION

   14.1  This method utilizes SPE to extract analytes from water. It requires the use of very
         small volumes of organic solvent and very small quantities of pure analytes, thereby
         minimizing potential hazards to both the analyst and the environment as compared to
         the use of large volumes of organic solvents in conventional liquid-liquid extractions.

   14.2  For information about pollution prevention that may be applicable to laboratory
         operations, consult "Less is Better: Guide to Minimizing Waste in Laboratories"
         available from the American Chemical Society's Department of Government Relations
         and Science Policy, 1155 16th Street N.W., Washington, D.C., 20036 or on-line at
         http ://portal. acs.org/portal/fileFetch/CAVPCP 012290/pdf/WPCP  012290.pdf
         (accessed June 2012).
                                        540-27

-------
15. WASTE MANAGEMENT

   15.1  Analytical procedures described in this method generate relatively small amounts of
         waste since only small amounts of reagents and solvents are used. The matrices of
         concern are finished drinking water or source water.  However, laboratory waste
         management practices must be conducted consistent with all applicable rules and
         regulations, and that laboratories protect the air, water, and land by minimizing and
         controlling all releases from fume hoods and bench operations.  Also, compliance is
         required with any sewage discharge permits and regulations, particularly the hazardous
         waste identification rules and land disposal restrictions.

16. REFERENCES

1.  Winslow, S.D., Pepich, B.V., Martin, J.J., Hallberg, G.R., Munch, D.J., Frebis, C.P.,
   Hedrick, E. J., Krop, R. A. "Statistical Procedures for Determination and Verification of
   Minimum Reporting Levels for Drinking Water Methods." Environ. Sci. Technol. 2004, 40,
   281-288.

2.  Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, W.L. Budde, "Trace Analyses for
   Wastewaters." Environ. Sci. Technol. 1981, 15, 1426-1435.

3.  Leenheer, J.A., Rostad, C.E., Gates, P.M., Furlong, E.T., Ferrer, I.  "Molecular Resolution
   and Fragmentation of Fulvic Acid by Electrospray lonization/Multistage Tandem Mass
   Spectrometry.'Mwa/.  Chem. 2001, 73, 1461-1471.

4.  Cahill, J.D., Furlong E.T., Burkhardt, M.R., Kolpin, D.,  Anderson, L.G.  "Determination of
   Pharmaceutical Compounds in Surface- and Ground-Water Samples by Solid-Phase
   Extraction and High-Performance Liquid Chromatography Electrospray lonization Mass
   Spectrometry." J. Chromatogr. A, 2004, 1041, 171-180.

5.  "Prudent Practices in the Laboratory: Handling and Disposal of Chemicals," National
   Academies Press (1995), available at http://www.nap.edu (accessed October 5, 2011).

6.  "OSHA Safety and Health Standards, General Industry," (29CFR1910), Occupational Safety
   and Health Administration, OSHA 2206, (Revised, July 2001).

7.  "Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
   Committee on Chemical Safety,  7th Edition. Available  online at
   http://portal.acs.org/portal/PublicWebSite/about/governance/committees/chemicalsafetv/publ
   ications/WPCP 012294 (accessed August 2013).

8.  Winslow, S. D. , Pepich, S. D. , Bassett, M. V., Wendelken, S. C., Munch, D. J.,  Sinclair, J.
   L. "Microbial inhibitors for U.S. EPA drinking water methods for the determination of
   organic compounds." Environ. Sci. Technol., 2001, 35,  4103-4110.
                                         540-28

-------
17. TABLES, DIAGRAMS, FLOWCHARTS AND VALIDATION DATA
TABLE 1. LC METHOD CONDITIONS
Time (min)
Initial
8.0
9.0
28.0
28.1
30.0
30.1
40.0
% Formate Buffer
90.0
60.0
50.0
17.7
10.0
10.0
90.0
90.0
% Methanol
10.0
40.0
50.0
82.3
90.0
90.0
10.0
10.0
Restek Ultra Aqueous 2. 1 x 100 mm packed with 5 jam Cig stationary phase
Flow rate of 0.3 mL/min
10 |aL partial loop injection into a 20 jiL loop
        TABLE 2. ESI-MS/MS METHOD CONDITIONS
ESI Conditions
Polarity
Capillary needle voltage
Cone gas flow
Nitrogen desolvation gas
Desolvation gas temp.
Positive ion
4kV
lOOL/hr
HOOL/hr
350 °C
                              540-29

-------
TABLE 3. METHOD ANALYTE SOURCE, RETENTION TIMES (RTs), AND
          SUGGESTED IS REFERENCES
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13Cz 15N(SUR)
tebuconazole-t4 (SUR)
carbofuran-13C6(IS#l)
bensulide-Ji4(IS#2)
phorate- dw (IS#3)
Method Analyte
Source8
Accu Standard
Absolute Standards
Absolute Standards
Absolute Standards
Spex CertiPrep
Crescent Chemical
Chem Service
Absolute Standards
Accu Standard
Chem Service
Crescent Chemical
Crescent Chemical
Cambridge Isotopes
Dr. Ehrenstorfer
Cambridge Isotopes
Cambridge Isotopes
Crescent Chemical
Peak#
(Fig. 1)
1
3
5
6
7
8
9
11
12
13
14
17
2
16
4
10
15
RT
(mill)
6.14
8.09
12.30
12.56
13.12
13.53
13.91
19.14
19.17
19.62
19.64
20.93
6.13
20.88
11.45
18.96
20.35
IS# Ref
1
1
1
1
3
O
3
2
2
O
2
2
1
2
_
_
-
  Data presented in this method were obtained using analytes purchased from these
  vendors. Other vendors' materials can be used provided the QC requirements in
  Section 9 can be met.
                              540-30

-------
      TABLE 4. MS/MS METHOD CONDITIONS
Segment11
1
2
2
2
2
2
2
O
3
O
3
O
1
O
2
O
3
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13C2,15N
tebuconazole-de
carbofuran-13Ce
bensulide-t/i4
phorate-t/io
Precursor
Ion c (m/z)
163
238
336
320
293
211
291
398
353
334
304
308
166
314
228
412
271
Product
Ionc'd (m/z)
88
163
266
233
171
97
185
314
133
278
217
70
91
72
171
316
75
Cone
Voltage (v)
15
20
30
30
20
20
20
20
15
25
30
30
15
30
26
20
15
Collision
Energy6 (v)
10
15
20
25
10
30
15
10
20
20
25
20
10
20
13
10
10
An LC/MS/MS chromatogram of the analytes is shown in Figure 1.
Segments are time durations in which single or multiple scan events occur.
Precursor and product ions listed in this table are nominal masses. During MS and MS/MS
optimization, the analyst should determine the precursor and product ion masses to one
decimal place by locating the apex of the mass spectral peak place (e.g., m/z 398.1^313.9 for
bensulide).  These precursor and product ion masses (with one decimal place) should be used
in the MS/MS method for all analyses.
Ions used for quantitation purposes.
Argon used as collision gas at a flow rate of 0.3 mL/min.
                                      540-31

-------
TABLE 5. DLs AND LCMRLs IN REAGENT WATER
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
Fortified
Cone. (ng/Lf
1.60
1.60
1.60
1.60
1.60
1.60
1.60
1.60
0.64
1.60
0.64
0.64
Oasis HLB
DLb(ng/L)
0.39
0.45
0.88
0.39
0.57
0.70
0.68
0.51
0.26
1.0
0.30
0.25
LCMRLC
(ng/L)
1.2
1.3
1.0
0.86
0.99
2.0
2.0
1.2
0.81
2.0
0.64
2.0
Speedisk H2O-Philic DVB
DLb(ng/L)
0.32
0.25
0.42
0.32
0.32
0.46
0.37
1.1
0.15
0.77
0.26
0.30
LCMRLC
(ng/L)
0.61
1.2
1.0
1.6
1.0
1.1
1.1
1.4
0.73
2.7
0.30
0.33
 Spiking concentration used to determine DL.
' Detection limits were determined by analyzing seven replicates over three days according to
 Section 9.2.6.
 LCMRLs were calculated according to the procedure in reference 1.
                                        540-32

-------
TABLE 6. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES IN
         REAGENT WATER FORTIFIED AT A HIGH CONCENTRATION (n=4)
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13C2,15N
tebuconazole-de
Fortified
Cone. (ng/L)
32.0
32.0
32.0
32.0
32.0
32.0
32.0
32.0
12.8
32.0
12.8
12.8
16.0
16.0
Oasis HLB
Mean %
Recovery
106
104
104
106
104
111
115
90.0
103
97.1
96.9
100
99.0
98.1
% RSD
6.3
2.6
4.1
4.5
5.1
2.6
4.4
4.3
5.6
5.6
2.7
5.2
4.3
8.6
Speedisk H2O-Philic DVB
Mean %
Recovery
105
102
98.4
102
111
116
117
92.7
95.5
104
96.7
97.6
99.8
92.6
% RSD
2.5
3.5
1.8
3.4
1.5
3.1
3.8
13
11
4.3
11
10
2.4
12
                                 540-33

-------
TABLE 7. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES IN
         REAGENT WATER FORTIFIED AT A LOW CONCENTRATION (n=4)
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13C2,15N
tebuconazole-de
Fortified
Cone. (ng/L)
4.00
4.00
4.00
4.00
4.00
4.00
4.00
4.00
1.60
4.00
1.60
1.60
16.0
16.0
Oasis HLB
Mean %
Recovery
102
101
101
108
94.2
106
113
93.6
110
95.5
92.6
115
101
98.9
% RSD
1.9
6.1
3.8
2.0
3.2
3.0
2.9
5.5
5.2
1.3
6.8
4.9
3.4
1.4
Speedisk H2O-Philic DVB
Mean %
Recovery
103
107
99.6
108
108
112
111
93.0
105
112
104
114
98.1
105
% RSD
2.6
4.8
3.1
9.6
7.2
4.3
6.5
3.0
2.7
11
6.6
3.4
2.1
4.8
                                 540-34

-------
TABLE 8. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES
         FORTIFIED AT A HIGH CONCENTRATION IN FINISHED DRINKING
         WATER FROM A GROUND WATER SOURCE8 (n=4)
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13C2,15N
tebuconazole-de
Fortified
Cone. (ng/L)
32.0
32.0
32.0
32.0
32.0
32.0
32.0
32.0
12.8
32.0
12.8
12.8
16.0
16.0
Oasis HLB
Mean %
Recovery
105
101
103
106
97.8
102
103
97.1
100
91.8
104
100
103
93.5
% RSD
1.3
2.3
4.4
3.2
4.9
4.2
4.3
5.9
4.2
3.5
4.5
5.2
2.3
7.5
Speedisk H2O-Philic DVB
Mean %
Recovery
109
107
99.0
105
103
104
108
97.1
102
95.1
103
101
105
98.6
% RSD
2.0
1.5
1.6
1.3
1.5
2.5
2.1
1.2
4.3
2.2
4.4
4.0
1.0
3.9
  TOC = <0.50 mg/L and hardness = 342 mg/L as calcium carbonate.
                                   540-35

-------
TABLE 9. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES
         FORTIFIED AT A LOW CONCENTRATION IN FINISHED DRINKING
         WATER FROM A GROUND WATER SOURCE8 (n=4)
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13C2,15N
tebuconazole-de
Fortified
Cone. (ng/L)
4.00
4.00
4.00
4.00
4.00
4.00
4.00
4.00
1.60
4.00
1.60
1.60
16.0
16.0
Oasis HLB
Mean %
Recovery
101
103
108
99.4
98.1
109
115
88.8
100
124
96.9
98.4
104
96.3
% RSD
4.3
5.4
6.1
3.2
2.4
4.0
1.8
17
5.1
14
3.7
3.2
3.0
5.5
Speedisk H2O-Philic DVB
Mean %
Recovery
103
107
107
100
100
108
111
85.6
90.6
102
98.4
85.9
101
91.7
% RSD
3.4
4.4
1.2
2.0
2.9
7.6
6.2
6.0
8.9
5.1
9.5
7.0
1.7
4.4
  TOC = 0.61 mg/L and hardness = 377 mg/L as calcium carbonate.
                                   540-36

-------
TABLE 10. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES
          FORTIFIED AT A HIGH CONCENTRATION IN FINISHED DRINKING
          WATER FROM A SURFACE WATER SOURCE8 (n=4)
Analyte
Methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
Bensulide
Tebufenozide
chlorpyrifos oxon
Fenamiphos
Tebuconazole
methomyl-13C2,15N
tebuconazole-de
Fortified
Cone. (ng/L)
32.0
32.0
32.0
32.0
32.0
32.0
32.0
32.0
12.8
32.0
12.8
12.8
16.0
16.0
Oasis HLB
Mean %
Recovery
106
101
106
108
102
105
110
101
101
102
107
107
100
98.9
% RSD
2.2
1.4
0.6
1.8
2.6
2.1
1.8
7.6
5.3
4.1
5.8
5.4
3.5
6.3
Speedisk H2O-PhiIic DVB
Mean %
Recovery
111
105
104
104
105
105
107
100
96.7
103
99.5
107
107
101
% RSD
2.5
1.3
2.5
0.7
6.0
5.2
6.0
7.5
6.1
8.0
6.8
7.5
1.8
6.4
  TOC = 1.22 mg/L and hardness =154 mg/L as calcium carbonate.
                                   540-37

-------
TABLE 11. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES
          FORTIFIED AT A LOW CONCENTRATION IN FINISHED DRINKING
          WATER FROM A SURFACE WATER SOURCE (n=4)
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-13C2,15N
tebuconazole-de
Fortified
Cone. (ng/L)
4.00
4.00
4.00
4.00
4.00
4.00
4.00
4.00
1.60
4.00
1.60
1.60
16.0
16.0
Oasis HLBa
Mean %
Recovery
104
96.6
97.6
106
90.3
102
111
90.4
98.9
94.0
96.1
111
101
100
% RSD
2.1
4.3
2.2
3.1
3.3
5.5
4.8
7.1
9.5
4.6
3.9
6.6
0.8
2.8
Speedisk H2O-Philic DVBb
Mean %
Recovery
109
106
104
105
98.1
105
107
96.4
105
104
100
120
103
101
% RSD
3.1
3.4
4.3
7.4
6.6
5.7
4.4
12
11
7.6
3.1
5.2
1.9
3.7
  TOC = 1.22 mg/L and hardness =154 mg/L as calcium carbonate.
 ' TOC = 1.58 mg/L and hardness =137 mg/L as calcium carbonate.
                                    540-38

-------
TABLE 12.  AQUEOUS SAMPLE HOLDING TIME DATA FOR SAMPLES OF FINISHED DRINKING WATER FROM A
           SURFACE WATER SOURCE3, FORTIFIED WITH METHOD ANALYTES AND PRESERVED AND STORED
           ACCORDING TO SECTION 8 (n=4)h
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-1JC2,i:'Nc
tebuconazole-de0
Fortified
Cone.
(ng/L)
32.0
32.0
32.0
32.0
32.0
32.0
32.0
32.0
12.8
32.0
12.8
12.8
16.0
16.0
DayO
Mean
%Rec
106
101
106
108
102
105
110
101
101
102
107
107
100
98.9
%
RSD
2.2
1.4
0.6
1.8
2.6
2.1
1.8
7.6
5.3
4.1
5.8
5.4
3.5
6.3
Day?
Mean
%Rec
104
95.2
103
103
95.9
100
108
99.2
104
79.6
98.4
104
101
91.7
%
RSD
2.5
2.8
1.5
3.0
5.2
5.2
5.0
5.5
7.4
6.2
4.1
6.9
2.5
8.2
Day 14
Mean
%Rec
109
94.7
100
103
89.1
93.2
101
100
109
62.4
104
110
105
102
%
RSD
1.3
5.1
3.2
3.5
11
11
11
4.1
5.8
13
4.4
7.2
1.7
5.4
Day 23
Mean
%Rec
109
90.5
99.4
105
101
105
112
106
108
59.1
107
108
102
102
%
RSD
3.8
1.4
2.1
2.4
1.2
1.5
3.6
6.0
4.1
3.0
5.5
7.1
1.5
6.8
Day 28
Mean
%Rec
113
93.9
104
108
100
103
110
100
101
56.0
98.7
106
104
95.5
%
RSD
1.6
3.6
4.3
1.4
2.8
3.5
5.1
4.6
5.0
3.7
5.9
2.6
2.6
6.4
  TOC = 1.22 mg/L and hardness =154 mg/L as calcium carbonate.
  Oasis HLB SPE cartridges used for holding time studies.
 0 Surrogates were not added to samples until the day of extraction.
                                                 540-39

-------
TABLE 13. EXTRACT HOLDING TIME DATA FOR SAMPLES OF FINISHED DRINKING WATER FROM A
          SURFACE WATER SOURCE3, FORTIFIED WITH METHOD ANALYTES AND PRESERVED AND
          STORED ACCORDING TO SECTION 8 (n=4)a
Analyte
methomyl
3 -hydroxycarbofuran
fenamiphos sulfone
fenamiphos sulfoxide
phorate sulfone
phorate sulfoxide
disulfoton sulfoxide
bensulide
tebufenozide
chlorpyrifos oxon
fenamiphos
tebuconazole
methomyl-1 JC2,1:>N
tebuconazole-de
Fortified
Cone.
(ng/L)
32.0
32.0
32.0
32.0
32.0
32.0
32.0
32.0
12.8
32.0
12.8
12.8
16.0
16.0
DayO
Mean
%Rec
106
101
106
108
102
105
110
101
101
102
107
107
100
98.9
%
RSD
2.2
1.4
0.6
1.8
2.6
2.1
1.8
7.6
5.3
4.1
5.8
5.4
3.5
6.3
Day 7
Mean
%Rec
105
99.0
101
103
98.8
103
108
102
107
97.7
106
104
103
94.4
%
RSD
0.9
1.5
1.6
1.8
6.7
7.6
6.5
1.6
2.0
8.0
4.0
2.2
0.9
4.0
Day 14
Mean
%Rec
109
106
103
106
90.1
94.1
98.9
99.2
107
86.2
105
106
105
94.4
%
RSD
1.9
2.1
2.5
1.2
2.5
4.1
5.5
2.0
2.1
4.1
1.5
2.3
1.6
2.1
Day 23
Mean
%Rec
104
99.8
101
106
95.5
101
104
105
109
92.1
107
111
106
99.2
%
RSD
4.3
2.0
2.3
2.8
3.5
2.3
3.1
5.1
4.6
3.4
3.5
4.9
3.8
4.1
Day 28
Mean
%Rec
108
105
103
109
101
102
106
102
105
97.7
109
115
107
96.8
%
RSD
1.5
2.3
1.2
0.5
2.8
3.0
3.8
4.3
3.8
5.9
3.9
3.6
2.0
2.7
 Oasis HLB SPE cartridges used for holding time studies.
                                            540-40

-------
TABLE 14. INITIAL DEMONSTRATION OF CAPABILITY QUALITY CONTROL REQUIREMENTS
Method
Reference
Sect. 9.2.1
and 9.3.1
Sect. 9.2.2
Sect. 9.2.3
Sect. 9.2.4
Sect. 9.2.5
and 9.3.8
Requirement
Initial Demonstration of
Low System Background
Initial Demonstration of
Precision (IDP)
Initial Demonstration of
Accuracy (IDA)
Minimum Reporting Limit
(MRL) Confirmation
Quality Control Sample
(QCS)
Specification and Frequency
Analyze LRB prior to any other IDC steps.
Analyze four to seven replicate LFBs fortified near
the midrange calibration concentration.
Calculate average recovery for replicates used in
IDP.
Fortify, extract and analyze seven replicate LFBs
at the proposed MRL concentration. Calculate the
Mean and the Half Range (HR). Confirm that the
upper and lower limits for the Prediction Interval
of Result (Upper PIR, and Lower PIR, Sect.
9.2.4.2) meet the recovery criteria.
Analyze a standard from a second source, as
part of IDC.
Acceptance Criteria
Demonstrate that all method analytes are below 1/3 the MRL
and that possible interferences from extraction media do not
prevent the identification and quantification of method
analytes.
%RSDmustbe <20%
Mean recovery + 30% of true value
Upper PIR < 150%
Lower PIR > 50%
Results must be within 70-130% of true value.
NOTE:  Table 14 is intended as an abbreviated summary of QC requirements provided as a convenience to the method user. Because the information has been
        abbreviated to fit the table format, there may be issues that need additional clarification, or areas where important additional information from the method text
        is needed. In all cases, the full text of the QC in Section 9 supersedes any missing or conflicting information in this table.
                                                                    540-41

-------
TABLE 15.  ONGOING QUALITY CONTROL REQUIREMENTS (SUMMARY)
Method
Reference
Sect. 8.1-
Sect. 8.4
Sect. 8.4
Sect. 9.3.1
Sect. 9.3.3
Sect. 9.3.4
Sect. 9.3.5
Sect. 9.3.6
Sect. 9.3.7
Sect. 9.3.8
Requirement
Sample Holding Time
Extract Holding Time
Laboratory Reagent Blank
(LRB)
Laboratory Fortified Blank
(LFB)
Internal Standard (IS)
Surrogate Standards
(SUR)
Laboratory Fortified
Sample Matrix (LFSM)
Laboratory Fortified
Sample Matrix Duplicate
(LFSMD) or
Field Duplicates (FD)
Quality Control Sample
(QCS)
Specification and Frequency
28 days with appropriate preservation and storage as
described in Sections 8.1-8.4 except chlorpyrifos
oxon (7 days).
28 days when stored at < 6 °C and protected from
light.
One LRB with each extraction batch of up to 20 Field
Samples.
One LFB is required for each extraction batch of up
to 20 Field Samples. Rotate the fortified
concentrations between low, medium, and high
amounts.
Internal standards, carboruran-13C6, bensulide-
-------
TABLE 15.   (Continued)
  Method
  Reference
   Requirement
        Specification and Frequency
                 Acceptance Criteria
    Sect. 10.2
  Initial Calibration
Use IS calibration technique to generate a linear or
 quadratic calibration curve for each analyte.  Use
 at least five standard concentrations.  Check the
  calibration curve as described in Sect. 10.2.7.
When each CAL standard is calculated as an unknown using
the calibration curve, the analyte results must be 70-130% of
 the true value for all except CAL standards < MRL, which
 must be 50-150% of the true value.  If this criterion is not
   met reanalyze the CAL standards, restrict the range of
   calibration, or select an alternate method of calibration.
    Sect. 9.3.2
     and Sect.
       10.3
Continuing Calibration
    Check (CCC)
 Verify initial calibration by analyzing a low level
  (at the MRL or below) CCC prior to analyzing
  samples. CCCs are then injected after every 10
 Field Samples and after the last sample, rotating
 concentrations to cover the calibrated range of the
                 instrument.
Recovery for each SUR must be within 70-130% of the true
 value in all CCCs. Each analyte fortified at a level < MRL
  must calculate to be within ± 50% of the true value. The
  calculated concentration of the method analytes in CCCs
     fortified at all other levels must be within ± 30%.
NOTE: Table 15 is intended as an abbreviated summary of QC requirements provided as a convenience to the method user.  Because the information has been
        abbreviated to fit the table format, there may be issues that need additional clarification, or areas where important additional information from the method text
        is needed.  In all cases, the full text of the QC in Section 8-10 supersedes any missing or conflicting information in this table.
                                                                        540-43

-------
FIGURE 1.  EXAMPLE CHROMATOGRAM (OVERLAID MS/MS SEGMENTS) OF A CALIBRATION STANDARD WITH
           METHOD 540 ANALYTES AT CONCENTRATION LEVELS OF 3.2-8.0 jig/L. NUMBERED PEAKS ARE
           IDENTIFIED IN TABLE 3.
 1001
  •e?
  S
  *
  <
  o -
                                                     S
                                                6
                                                      r
                                                     £»
                                                       f

                                                                                      161
400
6 BO
1.00
                                    10-00     1200     1400
                                                 tune,
1(5.00
IS,00
2000
                                                  540-44

-------