EVALUATION OF POLYURETHANE FOAM CARTRIDGES
        FOR MEASUREMENT OF POLYNUCLEAR
         AROMATIC HYDROCARBONS  IN AIR
C. C. Chuang, W. E. Bresler, and S. W. Hannan
        Battelle Columbus  Laboratories
            Columbus, Ohio  43201
          Contract  Number 68-02-3487
               Project Officer

               Nancy K. Wilson
  Methods Development and Analysis Division
 Environmental Monitoring Systems Laboratory
Research Triangle Park, North Carolina  27711
 ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
     OFFICE OF  RESEARCH  AND DEVELOPMENT
     U.S.  ENVIRONMENTAL PROTECTION AGENCY
RESEARCH TRIANGLE PARK, NORTH  CAROLINA  27711

-------
                                  DISCLAIMER

     The information  in  this document  has  been funded wholly  or in  part  by
the United  States Environmental  Protection  Agency under  Contract  68-02-3487
to Battelle Columbus Laboratories.   It  has  been subject to  the  Agency's  peer
and administrative review* and  it  has been  approved  for publication  as an EPA
document.  Mention of  trade  names or commercial products  does  not  constitute
endorsement or recommendation for use.

-------
                                   FOREWORD

     Measurement and monitoring  research  efforts  are  designated  to  anticipate
environmental  problems,  to support  regulatory  actions by  developing an  in-
depth  understanding  of the nature and processes  that impact health and  the
ecology,   to   provide   innovative  means  of   monitoring   compliance   with
regulations,  and  to evaluate  the effectiveness  of  health  and  environmental
protection  efforts  through  the  monitoring of  long-term  trends.    The  En-
vironmental  Monitoring  Systems   Laboratory,  Research  Triangle  Park,   North
Carolina,  has  responsibility  for  assessment   of  environmental   monitoring
technology  and  systems,  implementation   of agency-wide  quality   assurance
programs  for  air   pollution  measurement  systems,  and  supplying   technical
support  to  other   groups  in  the Agency  including  the Office of  Air  and
Radiation, the Office of Toxic Substances, and the Office  of Solid Waste.
     The  determination  of human  exposure  to toxic  organic  compounds  is  an
area of increasing  significance  to EPA.   The evaluation of polyurethane  foam
cartridges   for   polynuclear   aromatic   hydrocarbon   measurements   provides
important information that can be applied to the measurement of  the  extent  of
human exposure to  the polynuclear aromatic compounds.

                                Thomas  R,  Hauser,  Ph.D.
                                Director
                                Environmental Monitoring Systems  Laboratory
                                Research Triangle  Park,  North  Carolina  27711

-------
                                   ABSTRACT


     The  objective  of this  project was to  evaluate polyurethane  foam  (PUF)
cartridges as  collection  media  for quantification of vapor  phase  polynuclear
aromatic hydrocarbons (PAHs) in air.
     Two  cleanup  methods  for   PUF  cartridges   --  compression  rinsing  and
combined  compression  rinsing and  Soxhlet extraction  —  have  been  evaluated.
Both  methods  successfully  remove  interfering  material   and  background  PAHs
from  the  PUF.   The  compression  rinsing method  is  recommended because  it  is
easier, faster, and cheaper.
     Two  procedures   for  extraction  of PAHs  from  the   PUF matrix,  Soxhlet
extraction  and compression  rinsing,  were  compared.    Modified  EPA medium
volume  samplers  having  quartz   fiber filters to  collect  particles  and  PUF
cartridges to  trap vapors were  used.   Prior to  sampling,.known  quantities  of
perdeuterated  PAHs  were  spiked   into  each  cleaned PUF  cartridge.    Eight
samplers  were  operated  outdoors in  parallel  for 24  hours.    After  sampling,
four  PUF  samples were Soxhlet-extracted with 10 percent ether/hexane  and  the
other   four   PUF   samples  were   extracted   by   alternate  compression   and
decompression  fifty  times in the  same solvent.   These  sample  extracts  were
analyzed  by  on-column  injection,  electron  impact  gas  chromatography/mass
spectrometry   (El  GC/MS)  to  determine  PAHs.     The   results   showed   that
compression  rinsing   is  comparable to  conventional  Soxhlet extraction,  and
that  both methods  successfully remove  PAHs from  the   PUF  cartridges.    The
compression rinsing method was then used in the stability study.
     The  stability study  was carried out  to determine  the  stability  of PAHs
adsorbed on  PUF cartridges  as a function of storage time  between  collection
and extraction.  Two  sets of PUF  samples were collected for this  study.   The
first set of samples  was  stored  in the  presence  of light, and the  second  set
of  samples  was kept   in the  dark.   The storage  temperature for both  sets  of
samples  was  approximately  20°C,   The  samples  were  stored  for  one,   ten,
twenty,  or  thirty  days  and then  extracted  with  10  percent  ether/hexane.
Sample  extracts  were  analyzed   by El  GC/MS.    The levels  of  perdeuterated
benzo(a}pyrene decreased  significantly  during  storage.   The rate  of  decrease
was much  faster  when the PUF cartridges  were  stored  in  light.   Other  PAH
levels were not adversely influenced by the  storage  time.
     Selected  sample  extracts obtained  from  the  pilot study of  previous  work
(Task  35)  were solvent  exchanged   into dimethylsulfoxide  at  eight  different
concentration  levels.  These samples  were packed  in  dry  ice and sent  to  EPA,
HERL/ERC for microbioassay analysis.
     This report was  submitted  in  fulfillment of  Contract No,  68-02-3487  by
Battelle   Columbus    Laboratories   under   the    sponsorship   of   the   U.S.
Environmental  Protection  Agency.   This  report covers the  period of  August  1,
1984 to December  31,  1984, and work was  completed as  of December  31,  1984.

-------

-------
                                  CONTENTS
Foreword	     iii
Abstract ...... 	 ..........  	  .  .      iv
Tables	      vi
Abbreviations	,	    viii
Acknowledgment .  	 .................      ix

     1.   Introduction ...............  	  ....       1
     2.   Conclusions.	       3
     3,   Recommendations	       5
     4.   Experimental  Procedures	       7
     5.   Results and Discussions.	      15

References	      24

-------
                                   TABLES

Number                                                                  Page
  4,1  Level  of Deuterated  PAHs  In  the  spiking stock solution ....       9
  4.2  Level  of Non-deuterated PAHs in  the stock standard solution. .       9
  4.3  GC and HS operating  conditions	      12
  4.4  Mass distribution of sample  extracts  for microbioassay ....      14
  5.1  Levels of PAHs  in PUF cartridges cleaned by two
       different methods.  . 	  ...... 	 .....      16
  5.2  Recoveries of PAHs from spiked  PUF cartridges using two
       extraction methods—compression  and Soxhlet extraction ....      17
  5.3  Levels of PAHs  in PUF cartridges extracted by two
       methods—compression and  Soxhlet extraction. .	      17
  5.4  Recoveries of PAHs from PDF  cartridges spiked prior to
       sampling as a function of storage time.  Storage conditions:
       20°C,  in the light	        19
  5.5  Recoveries of PAHs from PUF  cartridges spiked prior to
       sampling as a function of storage time.  Storage conditions:
       20°C,  in the dark.  .	 .        20
  5.6  Levels of native  PAHs found  in  PUF cartridges as a function of
       storage time.  Storage conditions:  20°C, in the light ...        21
  5.7  Levels of native  PAHs found  in  PUF cartridges as a function of
       storage time.  Storage conditions:  20°C, in the dark. ...        22

-------
                            LIST OF ABBREVIATIONS

BaP         benzo(a}pyrene
CI          chemical ionization
Dg-PAH      native polynuclear aromatic hydrocarbons
DMSO        dimethyl sulfoxide
Dn-PAH      perdeuterated polynuclear aromatic hydrocarbons
El          electron impact
GC/MS       gas chromatography/mass spectrometry
PAC         polynuclear aromatic compounds
PAH         polynuclear aromatic hydrocarbons
PCB         polychlorinated biphenyls
RIC         recontructed ion chromatogram
PUF         polyurethane foam
                                     VII

-------
                                ACKNOWLEDGMENT

     The  financial  support of  the  U.S.  Environmental  Protection  Agency  and
the thoughtful discussion of  Dr,  Nancy  K.  Wilson are gratefully acknowledged.
Technical assistance  from Dr.  Ralph  Riggin and  Mr. James  E.  Howes,  Jr.  is
appreciated.
                                    viii

-------
                                SECTION 1
                               INTRODUCTION

     Polynuclear  aromatic  compounds  have  been  extensively studied  and
have received  increased  attention in studies of air  pollution  in  recent
years  because  some  of  these  compounds  are  highly  carcinogenic  or
mutagenic.   To understand the  extent of human exposure  to   polynuclear
aromatic compounds,  reliable sampling and analytical  methodology must be
established  for monitoring the  concentrations  of  these compounds  in air.
In  general,  the   analytical  methodology  is  well  developed,  but  the
sampling procedures  can often reduce the validity  of  analytical  results.
Several  studies   {1-4}   have   shown   that   the   three-   to   four-ring
polynuclear  aromatic hydrocarbons  (PAH) in  air  may be  mainly  in  the
vapor  phase  and  are not  retained  by filters because  of  volatilization.
A  wide  variety   of adsorbents   such   as   Tenax-GC,  XAD-2   resin  and
polyurethane foam  (PUF)  has been used to sample organic vapors  (5).   The
PUF  cartridge is  easy  to  handle   in  the   field  and  has good  airflow
characteristics',   it  has  been  successfully  used   for   collection   of
pesticide and polychlorinated biphenyl (PCB)  vapors (6).
     Recently, Battelle  conducted a  study   {7}  to collect  ambient  and
indoor air  PAH using  a sampler  with  a  quartz  fiber  filter  and  a  PUF
back-up trap.   It was observed  that the PUF  cartridges  changed  from  a
pale white  to  a  light yellow color  over a  one week  storage period.   It
is not known whether this color change  during  storage is  associated with
any change  in the PUF cartridge's  ability  to retain  PAHs.   In a  large
scale air monitoring program, placement of the samplers in the  field  and
the  return   of  filters   and  traps  after air collection  to   analytical
laboratories may  involve  several  weeks.   Therefore  it  is necessary  to
conduct a  study  to  assess the  stability  of  PAHs adsorbed on  the  PUF
cartridges  during  storage.

-------
     The  objective  of this  project was  to  evaluate PUF  cartridges  for
collection  and subsequent  analysis of vapor  phase PAH  in  air.    Two
studies were performed:
     (a)  Extraction  study  to  evaluate two  extraction  procedures,
          Soxhlet extraction  and  compression rinsing,  for  removal
          of PAH from the PUF matrix.
     (b)  Stability  study  to determine  the  stability of  PUF  traps
          by examining the measured  PAH concentration as  a function
          of storage time between collection and extraction.
     The   sample   extracts   were   analyzed  by   electron  impact   gas
chromatography/mass  spectrometry  (El   SC/MS)  to   determine  both  native
PAHs and spiked perdeuterated PAHs.
     In addition,  sample  extracts  obtained from  the range-finding  study
done  under  a   previous  work  assignment   [Task   35  (7)]  were  solvent
exchanged into  dimethyl sulfoxide  at eight different levels and  sent  to
EPA, HERL/ERC for microbioassay.

-------
                                SECTION 2
                               CONCLUSIONS

     Two  PUF  cleanup methods,  compression  rinsing and  combined  com-
pression  rinsing and  Soxhlet  extraction,  have been  evaluated  for their
ability  to  remove  interfering material  and  background  PAHs  from  PUF
cartridges.   The results  showed  that the  compression-only  technique  is
comparable  to the  combined compression  and  Soxhlet extraction  method.
Thus  the compression  method  is  recommended  for  use in the  future  for
time and cost  savings.
     The  results of the extraction  study  indicated  that levels  of both
native and  perdeuterated  PAHs  found  in the PUF cartridges  were   similar
using  either  Soxhlet  extraction  or  compression  rinsing.   It has  been
demonstrated  that both methods can successfully  remove  PAHs from  the  PUF
cartridges.   Since  significant time  and cost  savings can  be achieved  by
using compression rinsing, this method was used in the stability study.
     Generally,  good  recoveries  for  the  spiked  perdeuterated  PAHs  were
obtained  for  PUF samples  extracted   immediately  after   collection,  with
the  exception of  Da-naphthalene.    This  finding  demonstrated that  PUF
cartridges cannot quantitatively retain volatile  two-ring PAHs  under  the
sampling conditions employed.  Greater loss of volatile components would
be expected  at higher sampling temperatures.   It  should  be  noted  that
cyclopenta(c,d)pyrene  and  higher  molecular weight  (_>252)  PAHs were  not
detected in  the  PUF  samples.   1-Nitropyrene  was also  not  found   in  the
PUF samples.
     The stabilities   of  PAHs  adsorbed on  PUF cartridges during  storage
with and  without light were investigated.    The  results indicated  that
levels  of BaP decreased   significantly during  storage.   The  rate of
decrease  was  much   faster  when   PUF  cartridges  were  stored  in   the

-------
presence  of  light.    The  levels  of  the  remaining  PAHs  were  not
significantly influenced  by  the  storage time.

-------
                                SECTION 3
                             RECOMMENDATIONS

     Both  XAD-2  resin and polyurethane  foam  (PUF)  are commonly  used  to
collect  PAH  vapors  in ambient air sampling (8).  A comparative  study  is
recommended  to  evaluate  the  flow  characteristics  and  the  collection
efficiency for  PAH  vapors by using these two  adsorbents  (XAD-2  and  PUF)
as  backup traps  in ambient  air  sampling.    Recently, several  research
groups   indicated  that  extracts  of  clean  PUF  plugs  show mutagenic
activity  and  interfere with bioassay  results  (9).   Therefore,  bioassay
analysis  is  recommended for  samples  collected  from  these  two adsorbents.
The bioassay results  can  provide  information  as to whether  PUF  plugs  or
XAD-2  resin  would  interfere  with  bioassay.    To   characterize  and  to
compare  these two adsorbents  thoroughly,  studies are also  recommended  to
determine the stability of PAHs captured  on XAD-2 resin as  a function  of
time.
     In  this  study,   quartz  fiber  filters were   used  to  collect air
particulate matter.  However, there are other  types  of filters which can
be considered  for  collection  of  particles.   Very  few studies have  been
conducted  to evaluate  filter material  for  collection   of  particle-bound
PAHs.  Lee's group  (10) has  evaluated  various  candidate filter materials
such as  glass  fiber,  quartz  fiber,  microglass fiber with Teflon  binder
and  Teflon membrane  filters.   Experiments  were  performed by  liquid-
spiking  BaP  onto  filters to  determine  the  recoveries.     However, the
liquid-spiking BaP does not represent  the native adsorption  state  of BaP
in  ambient  air  sampling.    Therefore,  it   is   recommended   that  an
evaluation study be conducted to compare  different  types of  filters  such
as quartz  fiber,  glass fiber, and Teflon-coated  filters  for collection
of   PAHs  in   ambient   air   sampling.     Several   important    filter

-------
characteristics  such  as flow  characteristics  and collection  efficiency
should be addressed.
     Studies to  determine stabilities for different  types  of  filters  and
to determine  the effects of  storage time  on  PAH  samples collected  on
different types of filters  are  also recommended.

-------
                                SECTION 4
                         EXPERIMENTAL PROCEDURES

SAMPLING METHODOLOGY
      In  both  the extraction  and  stability  studies,  eight  EPA medium
volume  samplers  using  General  Metals  bypass motors  in  place  of  the
original high  volume  blower motors were employed.   Quartz fiber  filters
were  used   to  collect  particulate matter  and  polyurethane  foam (PUF)
cartridges  were  used  to collect  vapors.   Samplers  were  located  outside
Battelle away  from any exhaust openings or  heavy traffic.   Two sets of
four  samplers  were placed  in parallel approximately  two  feet apart.   In
each  set,  the  samplers were  placed about one  foot  apart.   A five   foot
long  exhaust  hose  was  attached  to each  sampler leading away  from  the
sampler inlet  to  avoid recirculating the  exhaust air to  the filter  and
the PUF trap.   Prior  to  sampling, the sampler pump was turned on  and the
flow  rate  was  measured  by  means  of  a  calibration head  and  a  U-tube
manometer.    Adjustments  were made  with  a control and a  bypass  valve to
obtain a flow  rate of 8 cfm.   All  the  clean  PUF traps were spiked  with
100 ]iL of  a  methylene chloride  solution  containing  nominally 50 ug/mL of
each of five perdeuterated PAHs.   Then the spiked PUF cartridges and the
clean filters  were  placed  in  the  sampling  heads.   Air was sampled for 24
hours.  After  24  hours of  sampling, a  final flow check was  conducted on
each  sampler,  and  all  PUF  cartridges  and  filters  were processed  for
transport to the laboratory for analysis.

CHEMICAL ANALYSIS METHODOLOGY
Po lj> uret n_a_ne__Fo_am Ca rtr i dge __C1 ejinuj^ Method s
     Two   methods   for   cleanup   of   PUF   cartridges   were   examined:
compression  rinsing  followed  by  Soxhlet  extraction,  and  compression

-------
rinsing only.   Two PUF plugs were  cleaned  by compression rinsing, which
consisted  of  placing  a  PUF  plug  in  a   beaker  containing  800 ml  of
toluene,  compressing  it and  allowing  it to  expand  50 times,  using the
base of a 250 ml graduate cylinder to compress  the  foam.   This  process
was  repeated  using  acetone.    The PUF  cartridges  were  then  Soxhlet-
extracted with acetone for  16  hours  and dried  in  a  vacuum oven  using
water aspirator  vacuum at  room  temperature.  Another  two  PUF cartridges
were cleaned by  compression  rinsing and  drying as  above,  but the  Soxhlet
extraction with  acetone was  omitted.   After drying,  the four cleaned PUF
cartridges were  Soxhlet-extracted with 10 percent  ether in hexane  for 16
hours.    The  extracts  were  concentrated   to  1  ml  and  the  internal
standard, 9-phenylanthracene, was added  to  give  a  final  concentration of
5 ug/mt.  These  PUF  extracts were analyzed  by El  GC/MS to determine the
background levels of PAHs found  in the  cleaned PUF  cartridges.
     Polyurethane  foam  cartridges  to be  used in the recovery study, the
extraction  methods comparison,  and the storage  stability  study  were
cleaned  by  compression  rinsing  with  toluene  and  acetone   followed  by
Soxhlet extraction with acetone and drying  as  above.   Following cleanup,
the  PUF  cartridges  were  wrapped  with  hexane-rinsed  aluminum foil  and
placed  in 32-ounce jars closed  with  Teflon-lined caps until  they  were
used.   Typically, all  the  PUF cartridges  were  cleaned within 24  hours
prior to sampling.
Method ofPreparationof Standard Solutions
     The  spiking  stock  solution  containing  five  Dn-PAHs was  prepared  in
methylene  chloride and  was  stored  in  a 100 ml  volumetric  flask  at  a
nominal  concentration  level  of  50 yg/mL for  each  compound.   The  actual
concentrations of this  spiking  solution are  listed  in Table 4.1.   Four
1 ml aliquots  of the spiking stock  solution  were  transferred  into  four
2 ml GC vials  for field use.   Prior to  sampling,  exactly 100  yL of the
spiking  solution was withdrawn  into  a  100 uL  syringe,  and the  spike was
injected approximately 1 inch deep into the  cleaned PUF cartridge.
     Another stock solution containing selected native  PAHs  was prepared
at   a   nominal   concentration    of   100 yg/mL   per   compound,      The
                                    8

-------
              TABLE 4.1.  LEVEL OF DEUTERATED PAHs IN THE
                          SPIKING STOCK SOLUTION
     Compound
                                         Concentration of Standard,
                                                   yg/mL
Dg-Naphthalene
D^o-Phenanthrene
Dig-Pyrene
Di2-Benzo(a}pyrene
                                                     55
                                                     51
                                                     56
                                                     48
                                                     58
concentrations in this  stock  solution are  given in  Table 4.2.     This
stock solution was  used to prepare  nominal  1,  5  and 10  yg/mL  standard
solutions which  also contained  the  spiked  Dn-PAHs  and  the  internal
standard, 9-phenylanthracene,  at a  constant  concentration  of 5  yg/mL.
              TABLE 4.2.  LEVEL OF NON-DEUTERATED PAHs  IN
                          THE STOCK STANDARD SOLUTION
     Compound
                                         Concentration  of Standard,
                                                   ug/mL
Phenanthrene
Fluoranthene
Pyrene
Benz(a)anthracene
Benzo(e)pyrene
Benzo(a)pyrene
Benzo(g,h,i )perylene
Coronene
1-Nitropyrene
                                                    119
                                                    124
                                                    106
                                                    102
                                                    113
                                                    105
                                                     93
                                                    109
                                                    103

-------
Method of Extrac11 on Study
     Eight  PUF  and   filter  samples  were  collected   in  the  extraction
study.   The filter samples  were  wrapped with  aluminum  foil, stored  at
room -temperature, and  were  not analyzed  in  this  program.   Four  PUF
samples were  Soxhlet-extracted  with 800 ml  10 percent ether/hexane  for
16  hours.   After  extraction,  the sample  extracts  were  concentrated  to
I ml and transferred into 2 ml sample vials.
     Another  four  PUF samples were  extracted  with  800 ml  of  10  percent
ether/hexane  using  the  compression  technique.   The  PUF  plug  was  placed
in  a  beaker  containing  800 ml  of 10  percent  ether/hexane  and  was
compressed  fifty times  using  the base  of a  250 ml   graduate  cylinder.
Then the  sample extract  was evaporated to  1  ml and   transferred  into  a
2 ml sample  vial.   Prior  to GC/MS  analysis,  the  internal   standard,  9-
phenylanthracene,  was  added to  all  the  sample  extracts  to  give  a
constant concentration of 5  yg/rnL.  The  levels  of Dg-PAHs and  Dn-PAHs  in
these PUF samples were determined using  El GC/MS.
Method of Stability Study
     Two sets  of sampling were  conducted  in  the  stability study.   The
first set of sampling employed eight EPA medium volume samplers,  but  the
motors of  two samplers  burned  out  during  sampling.   Because the  total
volumes  sampled  for  these  two  samples were  not  known,  only  six  PUF
samples were  available  for  the  study.   The  sampling was   repeated  one
week later to obtain the second set of samples.  After 24 hour sampling,
eight PUF samples were collected for the stability  study.
     Neither  the  first  nor the second  set  of  filter   samples were
analyzed in  this study.   Only PUF  samples were used in  the  stability
study, these  were  stored  under  different conditions.  The   first set  of
PUF samples was  stored  in the presence  of white fluorescent light  (room
lights) at  room  temperature;  the  wavelengths  and  intensities  of  the
light  were  not  measured.    The  second  set  of PUF samples  was  wrapped
individually  with  aluminum  foil  and   stored  in  the  dark  at  room
temperature.   Two  PUF  samples   from each set  were   extracted  with   10
percent ether/hexane  usingthe  compression  technique  immediately  after

                                   10

-------
sampling.   The remaining PDF samples were  stored  either in the light or
in the dark  for  ten-,  twenty-  or thirty-day intervals  and extracted with
10 percent  ether/hexane.   Sample extracts were  concentrated  to  1 ml for
GC/MS analysis.   The  internal  standard,  9-phenylanthracene, was  added to
all sample extracts at a level of 5 pg/mL prior to GC/MS analysis.
Method of Analysis by Gas Chroma tography/
MassSpectrometry(GC/MS)
     Electron  impact  gas chromatography/mass spectrometry  (El  GC/MS) in
the fyll mass  scan mode  was  employed  to  determine  Dg-PAHs and the spiked
Dn-PAHs in the PUF samples.   The instrumental  conditions used are listed
in  Table  4.3.   A  Finnigan  Model   4500  quadrupole  mass  spectrometer
equipped  with an EI/C1  source  was  employed.    The mass ranges  were
scanned  from  50 amu  to  450 amu  at  1 sec  scan  rate.    The ion  source
temperature  was  held  at 160°C  and  the  electron multiplier was  operated
at approximately 10^ gain.
     Gas  chromatography  employed an  Ultra  #2  fused   silica  capillary
column  (50 m  x 0.31 mm I.D.,  0.17 urn  film  thickness,   Hewlett-Packard
Co.) and methane  carrier gas.    The GC column  interfaced directly to the
MS ion  source.   With  on-column  injection,  the  sample  was  injected  at
45°C to  prevent  thermal degradation  of  thermally labile  compounds  such
as 1-nitropyrene.  The GC column  temperature was  held  at 45°C for 2 min,
programmed rapidly to  100°C  in 5 min, and  then  programmed  from  100°C to
320°C at 6°C/min.
     The  standard solutions  and  sample extracts  were  analyzed by  El
GC/MS.  Identification  of  the PAH compounds  in  the sample  extracts was
based  on   the correct  mass  spectrum and   the  correct  retention  time
relative  to  the  internal   standard  as  determined  from  the   standard
analyses.     From  the  standard  analyses, the  response  factor  for  each
compound  relative to  the  internal   standard  was  calculated  over  the
calibration range.   The following  equation shows  the factors  on  which
quantification was based:
                            r  - "y X  C T t; X  Fy
                            LX "    A1s x Rf
                                   11

-------
                TABLE  4.3.   SC AND  MS  OPERATING  CONDITIONS
Chromatography
Column:

Carrier Gas:
Injection Volume:
Injection Mode:

Temperature Program
Initial Column Temperature:
Initial Hold Time:
Program:

Final Hold Time:

Mass Spectrometer'
Instrument:
lonization;
Emission Current:
Scan:
Preamplifier:
HP Ultra #2 cross!inked 51 phenyl methyl
silicone 50 m x 0.31 mm, 0.17 pm film
thickness
CH4 flow rate at 60 c«3/sec at 250°C
1 uL
On-column at 45°C
45°C
2 minutes
45° to- 100°C in 5 rain, then
100°C (5 min) to 320°C at 6°C/min
10 minutes
Finnigan 4500 GC/MS
Electron impact at 70 eV
0.3 ma
50-450 amu; 1.0 sec/scan
10~8 amp/volt
                                   12

-------
where
     Cx  = Total pg of a target compound in the extract
     Ax  = Molecular Ion area of a target compound
     CTS = Concentration of the internal standard
     A-JS = Molecular ion area of the internal  standard
     Rf  = Response factor of a target compound
     Fv  = Final volume of sample extract.
Method of Preparation of Sample Extracts for
Hicrobioassay
     Since  the microbioassay  is  conducted  directly  in  the vials  con-
taining  the  sample extracts,  the  vials  must  be  clean  and  free  of
chemical  or  biological  interference.    Therefore,  prior  to  solvent
exchange, the vials to  be  used  for shipping and  bioassay were  placed  in
a 500°C oven heated overnight, and  the screw  caps were Soxhlet-extracted
with methane! for 16 hours.  Sample extracts  from the  pilot  study  of the
previous   work  (7)   were  diluted   with   methylene   chloride  to   a
concentration of  3.77  mg/mL.   A one  ml  aliquot  was  removed  from  each
sample  extract  for  further  dilution  to  1 mg/ml   and  the  remaining
portions  were   stored   in  the  dark  at  -70°C.    The  diluted   extracts
(1 mg/mL)  were  then  divided  into  16  sample  vials  at  eight  different
levels of  mass in  duplicate.   The  conditions  for  sample  division  are
summarized  in  Table 4.4.   The sample  extracts were evaporated to  near
dryness  under  a gentle  nitrogen  stream  and  2 uL of  dimethylsulfoxide
(DMSO) was  added  to each  sample.   The DMSO  samples  were  mixed with  a
Vortex mixer and evaporated  under  nitrogen for additional five minutes.
Then the sample vials were sealed with screw  caps and  immediately  stored
under dry  ice.  The samples  were  packed  in dry  ice and  sent  to EPA  at
HERL/ERC for microbioassay with  and without metabolic activation  (+S9,-
S9).
                                   13

-------
   TABLE 4.4.   MASS  DISTRIBUTION  OF  SAMPLE EXTRACTS FOR
               MICROBIOASSAY
 Level  of Mass
in Sample V1al»
      ug
   Size of
Syringe Used,
     yL
Number of Vials
 Amount of
DMSO added,
     PL
     1000
      500
      200
      100
       50
       20
       10
        5
    1000
     500
     250
     100
      50
      25
      10
      10
       2
       2
       2
       2
       2
       2
       2
       2
     2
     2
     2
     2
     2
     2
     2
     2
                               14

-------
                                SECTION 5
                          RESULTS  AND  DISCUSSION

     The  results of cleaning  the  PUF  cartridges  by two different methods
are  summarized  in  Table 5.1.  Only small  quantities  of naphthalene were
found in  the  PUF  cartridges  cleaned by either the  compression  method  or
the  combined  compression  and Soxhlet  extraction  method.   However,  as
shown in  Figures  1 and 2, an unknown  impurity  peak  which  is  about  90
percent of the  total ehromatographic  peaks  was  detected in  the clean PUF
cartridges after both of these two different cleaning methods.
     The  results of  the extraction study are presented  in  Table 5.2 and
5.3.   As shown  in Table  5.2,  generally  good  recoveries (>85  percent)
were  obtained for spiked  Dn-PAHs with  the exception  of Dg-naphthalene
using either  the Soxhlet extraction or the  compression method.   The loss
of   Dg-naphthalene  was  anticipated   since  this  compound   is  highly
volatile.   The  sampling  temperature  during this  experiment  was 0°C  to
10°C.   Even  greater  losses of this volatile compound  can be expected  at
higher sampling temperatures.
     As shown  in  Table 5.3,  the  levels  of native PAHs found  in  the PUF
cartridges were  very  similar  using  these  two  methods.  The  levels  of
PAHs found in the  PUF  cartridges,  expressed as  ng per cubic  meter of air
sampled,  ranged  from  0.10  ng/m^  to  29.39 ng/m^  and  0.17  ng/m^  to
28.41 ng/m3 using  the  Soxhlet-extraction  and  the compression  methods,
respectively.    These data  demonstrate that the compression technique  is
comparable  to   Soxhlet-extraction  in   removing  PAHs  from   the  PUF
cartridges.   Since significant time and  cost savings can be achieved  by
using  the compression  method,  this  method was  used  in the  stability
study.
     Only  volatile  and   semi-volatile   PAHs  were   found  in   the  PUF
cartridges.   Higher molecular weight  PAHs(^252), such  as benzo(a)pyrene
and   coronene,   and    1-nitropyrene   were  not   detected  in    the  PUF
                                   15

-------
100.0-1
                                                                 5054460.
  RIC_
                 500
                 8:20
1000
16:40
1500
25:00
2000
33:20
  I
2509 SCAN
41:40 TIME
       Figure 1.  The total  ion chromatogram of PUF blank  cleaned by  the  compression method.

-------
100.0-1
 RIC
                                                                3665910.
                500
                8:20
1000
16:40
1500
25:00
2009
33:20
  I
2508 SCAN
41:40 TIME
     Figure 2.  The  total  ion chromatogram of PUF blank  cleaned by the combined compression
                 and Soxhlet  extraction method.

-------
cartridges,  indicating  that  these  compounds are  predominantly retained
on the filter.
     TABLE 5.1.  LEVELS OF PAHs IN PUF CARTRIDGES CLEANED BY
                 TWO DIFFERENT METHODS
       Compound
                                        Method of  Cleaning
Compression Method
  Total  Mass/PUF
  Cartridge, \ig
Compression followed
     by Soxhlet
  Extraction Method
   Total  Mass/PUF
    Cartridge,  ug
Naphthalene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benz(a}anthracene
Chrysene
Cyclo pen ta(c,d)pyrene
1-Ni tropyrene
Benzo(e)pyrene
Benzo(a)pyrene
Benzo(g»h,i)perylene
Coronene
       0.07
        0.10
          (1)
(1)  Not detected.
                                   18

-------
      TABLE  5.2    RECOVERIES OF PAHs FROM SPIKED PUF  CARTRIDGES USING TWO EXTRACTION
                   METHODS - COMPRESSION AND SOXHLET EXTRACTION
Soxhlet Extraction
Compound Name
Dg-Naphthalene
Dio-Phenanthrene
DiQ-Pyrene
Dj2-Chrysene
Di2-Benzo(a)pyrene

No. 1
0.83
92.62
107.62
106.57
92.18
TABLE 5
No. 2
0.69
91.20
101.13
102.42
87.53
No. 3
0.41
73.21
104.18
102.89
99.15
.3. LEVELS OF
METHODS -
. % Recovery
No.
2
84
103
93
103
, 4
.86
.14
.79
.60
.90
PAHs IN PUF
COMPRESSION
Soxhlet Extraction, Total Mass,
Compound Name
Naphthalene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benz(a)anthracene
Chrysene
Cyclopenta(c,d)pyrene
l-N1tropyrene
Benzo(e)pyrene
Benzo( a)pyrene
Benzo(g,h,1 Jperylene
Coronene
No. 1
4.14
30.18
2.33
7.58
5.06
..(1)
0.15
--
—
—
—
--
—
No. 2
3.87
26.93
2.02
6.66
4.33
0.03
--
--
--
—
—
—
No. 3
3.87
29.69
2.06
7.09
4.23
0.06
—
—
--
—
--
--
No
6.
30.
1.
7.
3.
0.
-

-
-


, nq/m3
. 4
56
74
57
12
56
15
-
-
-
-
-
-
Average
Value
1.20
85.29
104.18
101.37
95.69
No. 1
0.70
83.56
113.62
111.65
103.08
CARTRIDGES
AND SOXHLET

Average
Value
4.61
29.39
2.00
7.11
4.30
0.10
--
--
—
--
--
—

No. 1
3.40
26.35
1.72
6.75
4.39
0.12
—
--
--
--
--
--
Compression Method, %
No. 2
0.58
81.12
105.91
106.56
96.19
No. 3
1.16
89.33
108.13
102.32
93.76
Recovery
No. 4
0.67
86.57
99.90
92.17
90.40

Average
Value
0.78
85.14
106.89
103.18
95.86
EXTRACTED BY TWO
EXTRACTION
Compression
No. 2
3.77
26.17
1.54
6.60
3.90
0.15
—
--
--
—
--
—
Method, Total
No. 3
4.08
26.17
1.69
6.29
3.96
0.09
--
—
--
--
--
—
Mass, ng/m3
No. 4
4.94
34.94
2.39
7.94
5.40
0.31
—
—
--
--
--
--

Average
Value
4.05
28.41
1.84
6.90
4.41
0.17
--
--
—
--
—
--
(1) Not detected

-------
     The   stability   study  was   conducted  to   determine  whether   a
significant  loss  of PAHs  captured  by PDF  cartridges  occurs during  the
storage  period.   Two sets of  PUF samples were  obtained for  this  study.
The first  set  of  PUF samples was stored  at room temperature (~20°C)  in
the light, and  the second set  of samples was stored at  room temperature
in  the  dark.   Average  recoveries  of the  perdeuterated PAHs  extracted
from the PUF samples are  given  in Tables  5.4  and  5.5.   Concentrations  of
native  PAHs  found in  the PUF  samples are  calculated  as ng/m^ and  are
given in Tables 5,6 and 5,7.
     Recoveries for  the  perdeuterated  PAHs  for  the  first  set  of  PUF
samples  (stored in the light)  ranged  from 0.6 percent  for Dg-naphthalene
to 102.3 percent  for Di2~Chrysene.  Low recoveries  were  obtained  for  Dg-
naphthalene, which  decreased from  2.6  percent  to  0.4  percent  after  20
days storage.  The low recovery and the decreasing  recovery trend  for  Dg
-naphthalene during storage are mainly due  to the volatilization of this
compound.   The storage time does not appear  to have adverse effects  on
the recoveries  of DjQ-phenanthrene,  D}Q-pyrene,  and  D^-chrysene.   "^ne
variations  of recovery for these compounds  were less than 15  percent  and
may be  due to  small  variations  in  sampling  and  analysis procedures.   It
should  be   noted   that  the  recoveries  of  D|2-benzo(a)pyrene  decreased
significantly  with  increased  storage time:;  the  recoveries  decreased
from 92.1  percent to  11.7  percent after  storage  for  20  days.   It  is
possible that  oxidation  or  other degradation reactions of D^-BaP  may
occur during storage.
     In the second set of  PUF  samples (stored in  the dark), the relative
recovery data  are  similar  to  those  obtained  with  the  first  set   of
samples.   But  the recovery  of  D^-BaP decreased  less  in the second  set
of samples than  in the  first  set.   The  recoveries  decreased from 95.4
percent to  44.3 percent after 20 days storage.  Even after  30 days, 29.5
percent of  the D^-BaP was recovered.   It appears that  the  PUF-adsorbed
Dj2-&aP is  more stable  in the absence  of light.
                                   20

-------
     TABLE 5.4.  RECOVERIES OF PAHs FROM PUF CARTRIDGES SPIKED
                 PRIOR TO SAMPLING AS A FUNCTION OF STORAGE TIME.
                 STORAGE CONDITIONS:  20°C» IN THE LIGHT
   Compound Name
                                      Percent Recovery
1
                                Storage Time Between Sampling
                                    and Extraction, Days
                                              10
20
                      2.6(2.4,2.9)(D    0.6(0.9,0.3)
                     87.8(85.5,90.1)    90.8(87.2,93.4)
                     98.6(95.7,101.5), 85.8(91.5,81.1)
                    102.3(103.2,101.4) 88.8(86.1,91.5)
Di2-Benzo(a)pyrene   92.1(94.6,89.6)    49.5(48.2,50.8)
Dg-Naphthalene
DjQ-Phenanthrene
DiQ-Pyrene
                             0.4(0.4,0.4)
                            84.1(86.8,81.4)
                            91.4(88.4,94.4)
                            91.7(96.8,86.5)
                            11.7(12.1,11.3)
(1)  The first number given is  the mean  of the  duplicate  samples;
     the second and third numbers are  the range  of  the  samples.
                                   21

-------
             TABLE 5.5,  RECOVERIES OF PAHs FROM PUF CARTRIDGES SPIKED
                         PRIOR TO SAMPLING AS A FUNCTION  OF  STORAGE  TIME.
                         STORAGE CONDITIONS:   20°C,  IN  THE DARK
   Compound  Name
                                              Percent Recovery
 Storage Time Between Sampling and Extraction, Days
1              10                20                30
D8-Naphthalene      1.2(1.3,1.1)U)     2.7(2.2,3.2)       1.1(0.8,1.4)      0.6(0.9,0.3)
DjQ-Phenanthracene   88.3(86.9,89.7)   88.5(95.1,81.9)    84.5(87.2,81.8)   85.1(87.6,84.6)

D10-Pyrene           97.6(100.1,95.1)  93.3(90.4,96.2)    87.1(80.4,93.8)   91.2(89.6,92.8)

Di2-Chrysene        100.0(97.2,102.8)  94.5(91.4,97.6}   101.1(103.9,98.3)  89.4(81.7,97.1)

D12-Benzo(a)Pyrene   95.4(94.1,96.6)   51.9(49.8,56.0}   44.3(42.2,46.4)    29.5(27.2,31.8)
(1)  The first number given is the mean  of  the  duplicate samples;
     the second and third numbers  are  the range of  the samples.
                                           22

-------
             TABLE 5.6.  LEVELS OF NATIVE PAHs FOUND IN PUF CARTRIDGES
                         AS A FUNCTION OF STORAGE TIME.
                         STORAGE CONDITIONS:  2Q°C, IN THE LIGHT
    Compound  Name
                                        Concentration, ng/m3(l)
    Storage Time Between Sampling and Extraction, Days
          1                  10                  20
Naphthalene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benz{a)anthraeene
Chrysene
Cyclo penta(c,d)pyrene
l-N1tropyrene
Benzo(e)pyrene
Benzo(a)pyrene
Benzo(g,h,i)perylene
Coronene
 6.37(4.95,7.79)(2)    3.94(4.43,3.45}      2.95(2.73,3.17)
36.06(33.45,38.67)
 0.61(0.71,0.51)
 7.39(7.12,7.66)
 5.46(5.43,5.49)
 0.22(0.16,0.28)
 0.79(0.89,0.69)
       -(3)
29.41(26.80,32.02)  29.25(28.12,30.38)
 0.70(0.77,0.63)
 6.33(5.78,6.88)
 6.02(6.63,5.41)
 0.20(0.25,0.15)
 0.66(0.69,0.63)
0.41(0.38,0.44)
6.79(7.30,6.28)
5.84(6.04,5.64)
0.23(0.34,0.12)
0.71(0.81,0.61)
(1)  Expressed as ng per cubic meter of air sampled.
(2)  The first number given is the mean of the duplicate samples;
     the second and third numbers are the range of the samples.
(3)  Not detected
                                        23

-------
           TABLE 5.7.  LEVELS OF NATIVE PAHs FOUND IN PUF CARTRIDGES AS A FUNCTION OF
                       STORAGE TIME.  STORAGE CONDITIONS:  2QOC, IN THE DARK
Compound Name
                                                    Concentration, ng/m^U)
               Storage Time Between Sampling and Extraction,  Days
                      10                  20                  30
Naphthalene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benz(a)anthracene
Chrysene
Cyclopenta(c,d)pyrene
1-Nitropyrene
Benzo(e)pyrene
Benzo(a)pyrene
Benzo(g,h,i)pery!ene
Coronene
 9.80(9.41,10.19)(2)   9.04(9.43,8.65)     8.27(8.26,8.28)      6.17(6.47,5.87)
24.57(23.66,25.48)
 1.87(2.18,1.56)
 5.88(5.81,5.95)
 6.13(6.12,6.14)
 0.18(0.19,0.17)
 0.19(0.18,0.19)
      -(3)
29.72(32.96,26.48)  22.32(24.39,20.25)  29.38(29.91,28.85)
 0.95(1.01,0.89)
 6.03(6.52,5.53)
 6.68(6.60,6.76)
 0.17(0.22,0.12)
 0.14(0.13,0.14)
0.67(0.70,0.04)
5.82(5.78,5.86)
4.89(4.53,5.25)
0.09(0.12,0.06)
0.20(0.15,0.25)
1.01(1.04,0.98)
6.61(6.54,6.68)
5.56(5.95,5.17)
0.16(0.15,0.17)
0.18(0.18,0.18)
(1)  Expressed as ng per cubic meter of air sampled.
(2)  The first number given is the mean of the duplicate samples; the second and
     third numbers are the range of the samples.
(3)  Not detected.

-------
     Similar recovery  trends were  observed  for  the native  PAHs,   The
storage time does  not significantly affect  the  levels of  phenanthrene,
pyrene, and  chrysene  found  in  the PUF  samples.   Slightly  decreasing
concentrations  were detected for naphthalene.  Similar decreasing levels
were observed  for  anthracene,  an isomer of  phenanthrene.   The  reactive
PAH, cyclopenta(c.d)pyrene,  higher-molecular weight  PAHs  (>252), and 1-
nitropyrene were not detected in the PUF samples.
                                   25

-------
                               REFERENCES
(1)  Cautreels,  W.,   and   VanCauwenberghe,   K.,  "Experiments  on   the
     Distribution  of  Organic  Pollutants  Between  Airborne   Participate
     Matter and  Corresponding  Gas  Phase",  Atmos.  Environ., 12:1133-1141,
     1978.

(2)  Thrane, K.  E.,  and  Mikalsen,  A., "High Volume Sampling  of  Airborne
     Polycyclic  Aromatic  Hydrocarbons  Using  Glass  Fiber   Filters   and
     Polyurethane Foam", Atmos. Environ.,  1.5:909-918,  1981.

(3)  Feng,  Y.,  and  Bidleman,  T.  F., "Influence  of Volatility  on  the
     Collection   of   Polycyclic   Aromatic   Hydrocarbon   Vapors   with
     Polyurethane Foam", Envir. Sci.  Techno!.,  18:330-333, 1984.

(4)  Yamasaki, H.,  Kuwata,  K.,  and  Miyamoto, H.,  "Effects  of Ambient
     Temperature    on    Aspects    of   Airborne   Polycyclic    Aromatic
     Hydrocarbons", Envir.  Sci. Techno!..  16:182-194,  1982.

(5)  White, C. M.,  Sharkey, A.  G.,  Lee,  M. L.,  and  Vassilaros, D.  I.,
     "Some  Analytical   Aspects  of  the   Quantitative  Determination  of
     Polynuclear  Aromatic  Hydrocarbons  in  Fugitive  Emissions from Coal
     Liquefaction Process.   In:    Polynuclear  Aromatic  Hydrocarbons,  P.
     W. Jones, and  P.   Leber,  Editors.    Ann  Arbor Science Publication,
     Inc.  Ann  Arbor, Michigan,  pp.  261-275, 1979.

(6)  Billings, W.  N.,  and Bidleman,  T.  F., "High  Volume  Collection  of
     Chorinated Hydrocarbons in Urban Air using Three Solid Adsorbents"s
     Atmos. Environ.. 17:383-391,  1983.

(7)  Chuang, C.  C., Mack,  G.  A.,  Koetz, J.  R.,  and  Petersen,  B.  A,,
     "Pilot Study  of  Sampling  and   Analysis  for  Polynuclear  Aromatic
     Compounds  in   Microenvironments",   Final   Report   for    U.S.  EPA,
     Contract  No. 68-02-3487  (WA35),  1984.

(8)  Keller, C.  D.s  and Bidleman,  T. F.,  "Collection of Airborne Poly-
     cyclic Aromatic Hydrocarbons and Other  Organics  with  a   Glass  Fiber
     Filter  Polyurethane  Foam  System",   Atmos.   Environ.,   1^:837-845,
     1984.

(9)  Alfheim,  I., and Lindskog, A., "A Comparison Between Different High
     Volume Sampling Systems  for Collection Ambient  Airborne Particles
     for Mutagenicity  Testing  and  for  Analysis of  Organic   Compounds",
     Sci.  Total Environ., 34:203-222,  1984.
                                   26

-------
(10)  Lee,  F.  S.,  Pierson, W.  R.f and  Ezike,  J.   "The Problem  of  PAH
     Degradation  During  Filter  Collection  of Airborne  Participates --  an
     Evaluation  of Several  Commonly Used Filter Media.  In:  Polynuclear
     Aromatic  Hydrocarbons",  A  Bjerseth  and   A.   J.  Dennis,  Editors,
     Battelle  Press,  Columbus, Ohio,  pp. 543-563, 1979.
                                   27

-------
                                   TECHNICAL REPORT DATA
                            (Please read Instructions on the reverse before completing)
 1. REPORT NO,
   EPA/600/4-85/055
4, TITLE AND SUBTITLE
 evaluation of Polyurethane Foam Cartridges For
 Measurement of Polynuclear Aromatic  Hydrocarbons
 In Air.
                                                           3, RECIPIENT'S ACCESSION NO,
                                                          5, REPORT DATE
                                                             August 1985
                                                          6. PERFORMING ORGANIZATION CODE
7. AUTHORSS)

 C.  C.  Chuang, W.  E. Bresler, and S. W.  Hannan
                                                           8. PERFORMING ORGANIZATION REPORT NO
9, PERFORMING ORGANIZATION NAME AND ADDRESS

 Battelle Columbus Laboratories
 505  King Avenue
 Columbus, Ohio  43201
                                                           10. PROGRAM ELEMENT NO.
                                                          11. CONTRACT/GRANT NO.

                                                            68-02-3487, Task  41
 12. SPONSORING AGENCY NAME AND ADDRESS
  EPA,  Office  of Research and Development
  Environmental  Monitoring Systems Laboratory
  Research  Triangle Park, NC  27711
                                                           13. TYPi OF REPORT AMD PERIOD COVERED
                                                             Final  8/1/84 -  12/31/84
                                                          14. SPONSORING AGENCY CODE

                                                            EPA/600/08
15. SUPPLEMENTAH-Y NOTES
                                                                                have   been
                                                                           and   background
                                                                           because  it  is
              objective  of  this  project  was  -to  evaluate  polyurethane   foam   (PUF)
cartridges  as  collection media  for  quantification of vapor  phase polynuclear aromatic
hydrocarbons  (PAHs)  in  air.   Two  cleanup methods for  PUF cartridges  --   compression
rinsing  and  combined  compression  rinsing  and  Soxhlet  extraction
evaluated.   Both  methods   successfully  remove  interfering material
PAHs  from  the  PUF.   The  compression  rinsing method   is  recommended
easier, faster, and  cheaper.
     Two  procedures  for extraction  of PAHs  from the  PUF matrix,  Soxhlet extraction
and  compression  rinsing,  were  compared.    These sample extracts  were   analyzed   by
on-column  injection, electron  impact  gas  chromatography/mass  spectrometry (El GC/MS)
to  determine  PAHs.    The  results  showed  that compression  rinsing  is comparable   to
conventional  Soxhlet extraction,   and that  both  methods  successfully   remove  PAHs
from  the  PUF  cartridges.   The  compression  rinsing  method  was  then used in the
stability study.
     The stability study was  carried out  to  determine  the  stability of PAHs adsorbed
on  PUF  cartridges  as a function of  storage  time between collection  and  extraction.
The  results  indicated  that the  levels of   the  spiked  perdeuterated  benzo(a)pyrene
decreased  significantly during storage.   The rate  of  decrease  was much,  faster when
the  PUF  cartridges  were  stored  in  light.    Other PAH levels  were  not "adversely
influenced hv the  storage  time. ^             	  	
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
                                             b.IDENTIFIERS/OPEN ENDED TERMS  C.  COSATI Field/Group
 8. DISTRIBUTION STATEMENT

       RELEASE  TO  PUBLIC
                                             19. SECURITY CLASS (This Report)

                                                 UNCLASSIFIED
21. NO. OF PAGES

     36
                                              20. SECURITY CLASS (This page)
                                                   UNCLASSIFIED
EPA Form 2220-1 (Re*. 4-77)
                      PREVIOUS EDI T!ON IS OBSOLET E

-------