United States
Environmental Protection
Agency
                       EPA 600/R-11/091 | September 2011 | www.epa.gov/ord
             High Throughput Determination
             of Tetramine in Drinking Water
             by Solid Phase Extraction
             and Isotope Dilution Gas
             Chromatography/Mass
             Spectrometry (GC/MS)
            VERSION 1
Office of Research and Development
National Homeland Security Research Center
Centers for Disease Control and Prevention
Atlanta, Georgia

-------
                                                           EPA/600/R-11/091
HIGH THROUGHPUT DETERMINATION OF TETRAMINE IN DRINKING WATER
        BY SOLID PHASE EXTRACTION AND ISOTOPE DILUTION GAS
           CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS)
                                Version 1.0
                              September 2011
                               Developed by:

                   Centers for Disease Control and Prevention
                               In Support of:

                     U.S. Environmental Protection Agency

-------
                                    Disclaimer
The U.S. Environmental Protection Agency (EPA) through its Office of Research and
Development funded and managed the research described herein under EPA IA# DW75-
92259701 with the Centers for Disease Control and Prevention.  It has been reviewed by the
Agency but does not necessarily reflect the Agency's views. No official endorsement should be
inferred. EPA does not endorse the purchase or sale of any commercial products or services.

Questions concerning this document or its application should be addressed to:

Erin Silvestri, MPH (EPA Project Officer)
U.S. Environmental Protection Agency
National Homeland Security Research Center
26 W. Martin Luther King Drive, MS NG16
Cincinnati, OH 45268
513-569-7619
Silvestri.Erin@epa.gov

Matthew Magnuson, PhD (EPA Technical Lead)
U.S. Environmental Protection Agency
National Homeland Security Research Center
26 W. Martin Luther King Drive, MS NG16
Cincinnati, OH 45268
513-569-7321
Magnuson.Matthew@epa.gov

Rudolph Johnson, PhD
Centers for Disease Control and Prevention
4770 Buford Highway
MS F-44
Atlanta, GA 30341
770-488-3543
Rmj6@cdc.gov

Jennifer Links,  PhD
Centers for Disease Control and Prevention
4770 Buford Highway
MS F-44
Atlanta, GA 30341
770-488-4311
idr9@cdc.gov
                                          11

-------
                                 Table of Contents
Disclaimer	ii
List of Tables	iii
Foreword	iv
List of Acronyms	v
Acknowledgments	vi
CHAPTERS
1.     Scope and Application 	1
2.     Summary of Method 	2
3.     Definitions	3
4.     Interferences	4
5.     Safety 	5
6.     Equipment and Supplies	6
7.     Reagents and Standards	9
8.     Sample Collection, Preservation, and Storage	13
9.     Quality Control 	16
10.    Calibration and Standardization	22
11.    Procedure	25
12.    Data Analysis and Calculations 	28
13.    Method Performance	29
14.    Pollution Prevention	31
15.    Waste Management	31
16.    References	16
                                   List of Tables

Table 6-1. Gas Chromatograph (GC) Parameters	8
Table 7-1. Calibration Standard Stock Solution Volumes 	12
Table 8-1. Recoveries of Tetramine in Preservatives over Time	14
Table 8-2. Preservative Concentrations and Purposes of Preservatives	15
Table 13-1. Method Performance	29
Table 13-2. Single Lab Precision and Accuracy Data	29
Table 13-3. Percent Recovery of Tetramine for Several Tap Water Matrices and Residual
Disinfectants	30
                                         in

-------
Foreword

The National Homeland Security Research Center (NHSRC), part of the U.S. Environmental
Protection Agency's (EPA's) Office of Research and Development, is focused on developing
and delivering scientifically sound, reliable, and responsive products. These products are
designed to address homeland security information gaps and research needs that support the
Agency's mission of protecting public health and the environment. A portion of NHSRC's
research is directed at decontamination of indoor surfaces, outdoor areas, and water
infrastructure.  This research is conducted as part of EPA's response to chemical, biological, and
radiological  (CBR) contamination incidents. NHSRC has been charged with delivering tools and
methodologies (e.g. sampling and analytical methods, sample collection protocols) that enable
the rapid characterization of indoor and outdoor areas, and water systems following terrorist
attacks, and  more broadly, natural and manmade disasters.

The Selected Analytical Methods for Environmental Remediation and Recovery (SAM), formerly
referred to as the Standardized Analytical Methods for Environmental Restoration Following
Homeland Security Events, is a compendium of methods that informs sample collection and
analysis during the response to an incident.  SAM can be used by public and private laboratories
which are analyzing a large  number of samples associated with chemical, biological,  or
radiological  contamination.  Even though some of the analytes in SAM already have  existing
analytical methods, others are in need of improvements that enhance analytical capability and
meet more rigorous performance criteria.  Furthermore, not all of the analytical methods listed in
the SAM document address all possible matrices (e.g., water, soil, air, glass) encountered in
sample collection following an incident.   Some of the analytical methods in SAM have been
verified in a single laboratory, but most still need to undergo verification with respect to  a
specific contaminant in association with a specific matrix.

The sampling and analytical procedure (SAP) presented herein, describes a single laboratory
developed method  for the high throughput determination of tetramethylene disulfotetramine in
drinking water by solid phase extraction and isotope dilution gas chromatography/mass
spectrometry.  Performance data for this method have been generated in a single lab but the
method has not been studied jointly or independently by multiple labs. This method,  which will
be included in the SAM, is expected to provide the Water Laboratory Alliance, as part of EPA's
Environmental Response Laboratory Network, with a more reliable and faster means of analyte
collection and measurement.

Jonathan Herrmann,
Director, National Homeland Security Research Center
                                           IV

-------
Acronyms
   CAS	chemical abstract service
   CCC  	continuing calibration check
   CDC   	Centers for Disease Control and Prevention
   CR 	confirmation ratio
   CTMDL 	Chemical Terrorism Methods Development Laboratory
   DL 	detection limit
   El  	electron ionization
   FD 	field duplicate
   GC	gas chromatography
   HRpir	half range for the predicted interval of results
   HPLC  	high performance liquid chromatography
   i.d	inside diameter
   IDC 	initial demonstration of capability
   IS  	 standard
   ISTD  	internal standard primary dilution standard
   LD50  	median lethal dose
   LFB 	laboratory fortified blank
   LFSM 	laboratory fortified sample matrix
   LRB 	laboratory reagent blank
   MRL  	minimum reporting level
   MS	mass spectrometer
   MSDS	Material Safety Data Sheet
   m/z	mass to charge ratio
   NHSRC	National Homeland Security Research Center
   OSHA	Occupational Safety and Health Administration
   PIR	mean prediction interval of result ± half range for the predicted interval of results
   QC 	quality control
   RPD	relative percent difference
   SAM	Selected Analytical Methods for Environmental Remediation and Recovery
   	Restoration Following Homeland Security Events
   SAP	Sampling and analytical procedure
   Sect	section
   SIM  	selective ion monitoring
   SPE 	solid phase extraction
   SS  	standard solution
   TETS  	tetramine
   TOC	Total Organic Carbon

-------
Acknowledgments

The following researchers were critical to the development of the method and preparation of the
procedure and are acknowledged:

Centers for Disease Control and Prevention, National Center for Environmental Health
      Rudy Johnson
      Jennifer Links
The following individuals served as members and technical advisors of the Project Team and are
acknowledged:

U.S. Environmental Protection Agency (EPA), Office of Research and Development (ORD),
 National Homeland Security Research Center (NHSRC)
      Matthew Magnuson (EPA Technical Lead)
      Sanjiv Shah
      Erin Silvestri (EPA Project Officer)
                                        VI

-------
 HIGH THROUGHPUT DETERMINATION OF TETRAMINE IN DRINKING WATER
1.      SCOPE AND APPLICATION

       1.1.   This is a single laboratory developed isotope dilution gas chromatography/mass
              spectrometer (GC/MS) method for the determination of tetramethylene
             disulfotetramine (tetramine, TETS, Chemical Abstract Services Registry Number
             80-12-6). This method, including QC requirements, is designed to support site
             specific clean-up goals of environmental restoration activities following a
             homeland security incident involving this analyte.

       1.2.   Significance: Although banned in the United States, an accidental tetramine
             poisoning has been reported in New York City and several intentional poisonings
             have been reported in other countries, primarily in China [1,2]. Low levels of
             exposure can be  deadly and the human oral LDso (median lethal dose) has been
             reported to be as low as 0.1 mg/kg [2, 3]. Because tetramine is an odorless,
             tasteless white powder that easily dissolves in water but not absorbed through
             skin, the most common route of tetramine exposure is by ingestion [2].
             Symptoms of mild tetramine poisoning may include headache, dizziness, fatigue,
             weakness, lethargy, nausea, vomiting, perioral paresthesias (numbness around the
             mouth), and anorexia while high levels of exposures are characterized by seizures,
             coma and death [2].  Symptoms may begin  0.5-13 hours post exposure [2].

       1.3.   The use of 96-well plates for the solid phase extraction (SPE) produces two key
             benefits. The 96-well plates allow for extensive automation of the method,
             thereby enabling high throughput of samples, as might be required during
             environmental restoration.  Additionally, the use of this format results in the
             ability to perform isotope dilution by enabling the economical addition of
             isotopically labeled tetramine  as an internal standard to the sample prior to
             extraction.

       1.4.   Isotopically labeled tetramine  is added equally to all unknowns, quality controls,
             and calibration standards. In addition to enabling accurate quantitation,
             isotopically labeled tetramine  also accounts for and resolves some of the QC
             issues surrounding analysis, including analysis efficiency and sample loss, in the
             intended use of this analyte. The overall QC approach utilizing quantitation and
             confirmation ions as well as an isotopically labeled analyte greatly increases
             confidence that tetramine, and not another molecule with similar fragmentation
             patterns, is being quantitated during analysis.

       1.5.   This method was adapted from one that was initially developed by the Centers for
             Disease Control  and Prevention (CDC), in the National Center for Environmental
             Health (NCEH),  Division of Laboratory Sciences (DLS), Emergency Response
             and Air Toxicants Branch, in the Chemical  Terrorism Methods Development
             Laboratory (CTMDL) for the  determination and quantitation of tetramine in

-------
              human urine.  For the adapted method, accuracy and precision data have been
              generated in reagent water, and in finished ground and surface waters that use
              chlorine and/or chloramine as residual disinfectants.

       1.6.    The QC approach in this method conforms to CTMDL standards for clinical
              samples, and is presented here in terms more familiar to drinking water
              laboratories. Methods developed by CTMDL are distributed to the CDC's
              laboratory network, and the QC approach included in these methods, while  single
              lab verified by the CTMDL lab, is designed to be sufficiently rigorous that
              network labs can successfully perform the method.

       1.7.    The minimum reporting level (MRL) is the lowest analyte concentration that
              meets data quality objectives for the intended use of the method, e.g., to meet site-
              specific remediation goals. Laboratories will need to demonstrate that their
              laboratory MRL meets the requirements described in  Section 9.2.4.

       1.8.    Determining the detection limit (DL) is optional (Sect. 9.2.6).  Detection limit is
              defined as the statistically calculated minimum concentration that can be
              measured with 99% confidence that the reported value is greater than zero.

       1.9.    This method is intended for use by analysts skilled in the performance of solid
              phase extractions, the operation of GC/MS instruments, and the interpretation of
              the associated data.

       1.10.   This method has been verified using only the conditions and equipment specified
              in the method. Alteration of this method is not recommended.
2.     SUMMARY OF METHOD

       2.1.    A 50-mL water sample is collected, and a preservative and/or dechlorinating
              agent are optionally added as required by site-specific conditions. (The data in
              Table 8.1 suggest that the presence or choice of the additive does not affect the
              results.) An aliquot is  pipetted into a well of a preconditioned 96-well solid phase
              extraction plate, and the isotopically labeled tetramine is added.  Following a
              wash step, tetramine is then eluted in acetonitrile. The extract is concentrated to
              dryness under nitrogen and heat, and then adjusted to a 100 uL volume in
              acetonitrile. Tetramine is separated from the sample matrix and identified by
              GC/MS analysis, operated in SIM mode or equivalent. Analyte identification is
              accomplished by comparing the acquired mass spectra, including ion ratios, and
              retention times to reference spectra and retention times for calibration standards
              acquired under identical GC/MS conditions. Quantitation is performed using the
              internal standard technique. Utilization of an isotopically labeled internal standard
              provides a high degree of accuracy and precision for sample quantitation by
              accounting for analyte recovery and analytical efficiency.

-------
       2.2.   Compared to some drinking water methods (e.g. certain EPA 500 series methods),
             the initial laboratory demonstration of capability (IDC) is lengthier than some
             drinking water methods, the frequency of the on-going calibration is shorter, and
             the number of continuing calibration checks (CCC) is higher. Based on
             experience in the developer's lab, this QC approach ensures successful long-term
             implementation of the method in other labs, particularly when these methods are
             used infrequently (e.g. in emergency situations). Due to site-specific
             circumstances during an environmental remediation activity, e.g. in which sample
             throughput requirements exceed available lab capacity, a shorter initial
             demonstration of capability (IDC), changes to the on-going calibration frequency,
             and number of CCCs may be necessary and appropriate.. However,  initial and
             ongoing QC requirements and acceptance criteria (see Section 9) should not be
             changed. Adopting steps, such as a replacing on-going recalibration with a
             calibration check only, to save time may result in higher QC failure rates and
             perhaps less accurate quantitation. Labs should discuss these increased risks with
             sample submitters before taking such steps.

3.      DEFINITIONS

       3.1.   ANALYSIS BATCH - a sequence of samples, analyzed within a 24-hour period,
             including no more than 20 field samples in addition to all of the required QC
             samples (Sect. 9.3)

       3.2.   CALIBRATION STANDARD (CAL) - A solution prepared from the primary
             dilution standard solution and/or stock standard solution and the internal standard.
             The CAL solutions are used to calibrate the instrument response with respect to
             analyte concentration.

       3.3.   CONFIRMATION ION - for this method, the second most abundant tetramine
             ion (See Confirmation Ratio,  Sect. 3.4, below). The confirmation ion is used to
             calculate the confirmation ratio (Sect. 3.4)

       3.4.   CONFIRMATION RATIO (CR) - peak area produced by the confirmation ion
             divided by the peak area produced by the quantitation ion which serves as an
             additional QC measure of analyte selectivity

       3.5.   CONTINUING CALIBRATION CHECK (CCC) SOLUTION - a calibration
             solution containing the method analyte(s), which is extracted in the  same manner
             as the samples and analyzed periodically to verify the accuracy of the existing
             calibration for those analyte(s)

       3.6.   DETECTION LIMIT (DL) - the minimum concentration of an analyte that can be
             identified, measured, and reported to be greater than zero with 99% confidence

       3.7.   FIELD DUPLICATES  (FD1  and FD2) - two separate samples collected at the
             same time and place under identical circumstances, and treated exactly the same

-------
             throughout field and laboratory procedures to provide check the precision
             associated with sample collection, preservation, storage, and laboratory
             procedures

       3.8.   ISOTOPICALLY-LABELED INTERNAL STANDARD - a pure chemical added
             to an extract or to a standard solution in a known amount(s) and used to measure
             the relative response of other method analytes and surrogates that are components
             of the same solution.

       3.9.   LABORATORY FORTIFIED BLANK (LFB) - a volume of reagent water or
             other blank matrix to which known quantities of the method analytes and all the
             preservation reagents are added in the laboratory (Sect. 7.3.5.2)  The LFB is
             analyzed exactly like a sample and its purpose is to determine whether the
             methodology is in  control and whether the laboratory is capable of making
             accurate and precise measurements.

       3.10.  LABORATORY REAGENT BLANK (LRB) - an aliquot of reagent water that is
             treated exactly as a sample and used to determine if method analytes or other
             interferences are present in the laboratory environment, the reagents, or the
             apparatus

       3.11.  MATERIAL SAFETY DATA SHEET (MSDS) - written information provided
             by vendors concerning a chemical's toxicity, health  hazards, physical properties,
             fire, and reactivity data including storage, spill, and  handling precautions

       3.12.  MINIMUM REPORTING LEVEL (MRL) - the minimum concentration qualified
             to be reported as a quantitated value for a method analyte in a sample following
             analysis (Sect. 9.2.4. for MRL verification procedure)

       3.13.  PRIMARY DILUTION STANDARD SOLUTION - a solution containing the
             analytes prepared in the laboratory from stock standard solutions and diluted as
             needed to prepare calibration solutions and other needed analyte solutions

       3.14.  QUANTITATION ION -for this method, the quantitation ion is the parent
             tetramine ion with a mass to charge ratio (m/z) of 240 (See Confirmation Ratio,
             Sect. 3.4, above)

       3.15.  SECOND SOURCE QUALITY CONTROL SAMPLES - materials obtained
             from a source different than the original and used to verify the accuracy of the
             existing calibration for those analytes
4.     INTERFERENCES

      4.1    Method interferences that can lead to discrete artifacts and/or elevated baselines
             in the chromatograms may be caused by contaminants in solvents, reagents

-------
              (including reagent water), sample bottles and caps, and other sample processing
              hardware. All such items must be routinely demonstrated to be free from
              interferences under the conditions of the analysis by analyzing laboratory reagent
              blanks. Subtracting blank values from sample results is not permitted.

       4.2    Matrix interferences may be caused by contaminants that are co-extracted from
              the sample. The extent of matrix interferences will vary considerably from source
              to source, depending upon the nature of the water.

       4.3    Relatively high concentrations, in the mg/L range, of preservatives, antimicrobial
              agents, or dechlorinating agents might be added to sample collection vessels
              (Section 8.1.2). The  potential exists for trace-level organic contaminants in these
              reagents. Interferences from these sources should be monitored by analysis of
              laboratory reagent blanks particularly when new lots of reagents are acquired.

       4.4    Due to the nature of the matrix  analyzed in this procedure, occasional
              interferences from unknown substances might be encountered. Interfering
              compounds can be recognized by deviations in the sample
              quantitation/conformation ratios from the calibration standard ratios and can also
              be monitored using appropriate LRBs. Any interference that results in QC failure
              (Sect. 9) results in rejection of the entire analysis batch. If repeating the analysis
              does not remove the interference with the reference standard, the results for that
              analyte are not reportable.

       4.5    All glassware should be chemically cleaned before running this method. Wash
              glassware thoroughly with reagent-grade water followed by acetonitrile. Allow
              glass to dry completely before use. If the laboratory wishes to use a muffle oven
              for decontamination then the appropriate measures should be taken to assure that
              the muffle oven conditions are suitable to remove all traces of tetramine and other
              interferences.

       4.6    Care should be taken at all times to prevent contamination of QC materials,
              standards,  and samples.

       4.7    Chromatographic separation of the analyte should be carefully monitored for
              unknown interferences. See Section 11.2.5 for analyte confirmation.
5.      SAFETY
       5.1     The toxicity or carcinogenicity of each reagent used in this method has not been
              precisely defined. Each chemical should be treated as a potential health hazard,
              and exposure to these chemicals should be minimized.  Each laboratory is
              responsible for maintaining an awareness of OSHA regulations regarding safe
              handling of chemicals used in this method.  A reference file of MSDSs should be

-------
             made available to all personnel involved in the chemical analyses.  Additional
             references to laboratory safety are available [4-6].

       5.2    Tetramine is highly toxic, and the human oral LDso has been reported to be as low
             as 0.1 mg/kg [2, 3]. Ingestion is the primary reported route of exposure, but all
             other routes of exposure (e.g. inhalation, dermal contact, and eye exposure)
             should be avoided.  Follow universal safety precautions when performing this
             procedure, including the use of a lab coat, safety glasses, appropriate gloves, and
             a high quality-ventilated chemical fume hood and/or biological  safety cabinet.

       5.3    Avoid inhalation or dermal exposure to acetonitrile, which is used in the sample
             preparation steps.
       5.4    Mechanical hazards when performing this procedure using standard safety
             practices are minimal. Read and follow the manufacturer's information regarding
             safe operation of the equipment. Avoid direct contact with the mechanical and
             electronic components of the gas chromatograph and mass spectrometer, unless
             all power to the instrument is off. Generally, maintenance and repair of
             mechanical and electronic components  should be performed only by qualified
             technicians.
6.     EQUIPMENT AND SUPPLIES (It is important to note that specific brands or catalog
      numbers included in this section are examples only and do not imply endorsement of
      these particular products.  These specific products were used during the validation of this
      method.)

      6.1  MICRODISPENSERS - with adjustable volume (5-100 uL, 100-1000 uL)
           (Eppendorf Co., Westbury, NY or equivalent)

      6.2  REPEATER PIPETTE - 4780 (Eppendorf Co., Westbury, NY or equivalent)

      6.3  CONICAL AUTOSAMPLER VIALS - 300-uL vials (must be compatible with GC
           autosampler) especially for use as an alternative to 96 well plates

      6.4  ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g

      6.5   SOLID PHASE EXTRACTION (SPE) APPARATUS WITH 96 WELL PLATES

             6.5.1   96-WELL SPE PLATE - Strata X 60-mg / 6-mL (PN# 8E-S100-UGB),
                    available from Phenomenex (Torrence, CA) or equivalent

             6.5.2   PLATE SHAKER (ThermoFisher Scientific, Waltham, MA or equivalent)

-------
      6.5.3   96-WELL LIQUID HANDLER - Use a 96-well liquid handler equipped
             with a solid-phase extraction manifold and vacuum system. These
             systems must be calibrated prior to use, according to vendor or laboratory
             specifications. In addition, these liquid handlers must be used during
             laboratory-method validation.  The liquid handlers that have been used
             with this method in different laboratories have included the Tomtec®
             Quadra 3  SPE (Tomtec, Inc. Hamden, CT), the Caliper Zephyr, and the
             Caliper ilOOO (Caliper Life Sciences, Hopkinton, MA).  The selection of
             these liquid handlers has typically been based on cost and required sample
             throughput.

6.6  96-WELL NUNC DEEP WELL PLATE - 2000 mL plate (Nunc PN# 278752 or
     equivalent). Must be compatible with 96-well liquid handler described in Section
     6.5.3.

6.7  EXTRACT CONCENTRATION SYSTEM.  The 96-well plate requires a
     compatible dry-down step for sample pre-concentration following extraction.  The
     TurboVap 96 concentrator evaporator workstation (Zymark® Corp., Hopkinton,
     MA) has proven to be well suited for this application, but other evaporator systems
     which result in equivalent method performance could be used instead.

6.8  GAS CHROMATOGRAPHY ELECTRON IONIZATION MASS
     SPECTROMETRY SYSTEM (GC/MS)

     6.8.1  GC COLUMN - 30 m x 0.25-mm inside diameter (i.d.) fused silica capillary
           (5%-Phenyl)-methylpolysiloxane column coated with a 0.25um bonded film
           (Agilent HP-5ms [Agilent Technologies, Santa Clara, CA] or equivalent). A
           nonpolar, low-bleed column designed for GC/MS applications is
           recommended for use with this method to provide adequate chromatography
           and minimize column bleed.

     6.8.2  GC SYSTEM - The GC system  (e.g., Agilent 6890N GC or equivalent)
           must be equipped with an autosampler and injector and must provide
           consistent sample injection volumes. The system  should also be capable of
           performing linear temperature gradients at a constant flow rate.  The GC
           should be capable  of being configured exactly as stated below:

-------
Table 6-1. Gas Chromatograph (GC) Parameters
Parameter
GC Method
Column type
Injection Volume
Inlet liner
Inlet Temperature
Injection mode
Autosampler Tray
Temperature
Oven Program
Typical retention time
MS Scan Mode
lonization Type
Dwell Time
Setting
Constant flow at 1 mL/min
Initial pressure: 10.5 psi
Carrier Gas: Helium
HP5-ms (5% phenyl methyl siloxane), 30 m x 0.25 mm x 0.25 um
1 jiL
Splitless liner double taper, unpacked
250°C
Splitless injection; purge flow to split vent 100 mL/min at 1 min; gas saver
at 20 mL/min at 3 min
Room temperature
Initial temperature 100°C
Ramp 8°C/min to 200°C
Ramp50°C/minto300°C
Hold 3 00°C for 1.7 min
Tetramine =11.6 min
Selected ion monitoring (SIM)
Electron ionization (El)
100 msec per ion
             MASS SPECTROMETER (MS) - The MS (Agilent 5973 Mass Selective
             Detector, Palo Alto, CA, or equivalent) must be capable of performing electron
             impact ionization with positive ion detection and must be configured for selected
             ion monitoring (SIM, or equivalent depending on MS type) with a dwell time of
             100 msec per ion. The SIM ions monitored for this method should be set exactly
             as stated below:

                     Analyte                       Ion  (m/z)
                     Tetramine quantification ion     240

                     Tetramine confirmation ion      212

                     Tetramine internal standard      244

-------
7.     REAGENTS AND STANDARDS (These reagents were used during the validation of
      the method, and only these or their equivalent are acceptable for use.)

7. 1    GASES, REAGENTS, AND SOLVENTS - Reagent grade or better chemicals should be
      used. Unless otherwise indicated, it is intended that all reagents shall conform to the
      specifications of the Committee on Analytical Reagents of the American Chemical
      Society, where such specifications are available. Other grades may be used, provided it is
      first determined that the reagent is of sufficiently high purity to permit its use without
      lessening the quality of the determination.

      7.1.1  HELIUM - 99.9999% pure  or better, GC carrier gas

      7. 1 .2  REAGENT WATER - purified, deionized water which does not contain any
             measurable quantities of the method analyte or interfering compounds ([Tedia,
             Fairfield, OH] ® HPLC or equivalent grade water)

      7. 1 .3  METHANOL - (CH3OH, CAS#: 67-56-1) - high purity, demonstrated to be free
             of analytes and interferences (Tedia HPLC or equivalent)

      7. 1 .4  ACETONITRILE - (CH3CN, CAS#: 75-05-8) - high purity, demonstrated to be
             free of analytes and interferences (Tedia HPLC or equivalent)

      7.1.5  SAMPLE PRESERVATION REAGENTS - One of the following sample
             preservation reagents may be required by site specific conditions:

             7.1.5.1 AMMONIUM CHLORIDE (NH4C1, CAS#: 12125-02-9) - an additive
                   used in sample collection (Sigma-Aldrich ACS grade or equivalent)

             7.1.5.2 SODIUM THIOSULFATE (Na2S2O3, CAS#: 7772-98-7) - an additive
                   used in sample collection (Sigma-Aldrich ACS grade or equivalent)

             7.1.5.3 SODIUM SULFITE (Na2SO3, CAS#: 7757-83-7) - an additive used in
                   sample collection (Sigma-Aldrich ACS grade or equivalent)

             7.1.5.4 ASCORBIC ACID (C6H8O6, CAS#: 50-81-7) - an additive used in sample
                   collection (Sigma-Aldrich ACS grade or equivalent)

             7.1.5.5 AMMONIUM ACETATE (CH3CO2NH4, CAS#: 631-61-8) - An additive
                   used in sample collection (Sigma-Aldrich ACS grade or equivalent)

             7.1.5.6 CITRIC ACID (HOC(COOH)(CH2COOH)2, CAS#: 77-92-9) - an
                   additive used in  sample collection (Sigma-Aldrich ACS grade or
                   equivalent)
             7.1.5.7 DIAZOLIDINYL UREA (CgH^N^T, CAS#: 78491-02-8) - an additive
                   used for sample collection (Sigma-Aldrich ACS grade or equivalent)

-------
7.2   REAGENT PREPARATION

      7.2.1  5% METHANOL IN WATER - A 5%/95% methanol/water solution is prepared
             through volumetric dilution with HPLC grade deionized water. Measure 10 mL
             of methanol using an appropriate pipette, volumetric flask, or graduated cylinder
             and pour into a clean, dry container with a capacity of 250 mL or more.  Measure
             190 mL of HPLC grade deionized water with a volumetric flask or graduated
             cylinder and pour into the same container with the methanol. Mix the solution
             well.

7.3   STANDARDS SOLUTIONS - When a compound purity is assayed to be 96% or
      greater, the weight can be used without correction to calculate the concentration of the
      stock standard. Solution concentrations listed in this section were used to develop this
      method and are included as an example. Standards for sample fortification generally
      should be prepared in the smallest volume that can be accurately measured to minimize
      the addition of excess organic  solvent to aqueous samples. Store all calibration and
      control materials at either -20±5°C when not in use. Even though stability times for
      standard solutions  are suggested in the following sections, laboratories should use
      standard QC practices to determine when their standards need to be replaced.

      7.3.1  ISOTOPICALLY LABELLED INTERNAL STANDARD SOLUTIONS The
             internal standard used in this method is 13C4-tetramine (Cambridge Isotopes, MA;
             catalog #CLM-8146-0). Note that in  this method, the internal standard is a
             chemical that is structurally identical to the method analyte, but is substituted with
             13C. The isotopically-labeled internal standard has no potential to be present in
             water samples, and is not a method analyte. The internal standard is added to  all
             samples, standards, and QC solutions as described in Section 11.1.3.

      7.3.2  Prepare or purchase the internal standard at a concentration of 500 ng/mL. Steps
             for the preparation of this mixture  are described below:

             7.3.2.1 INTERNAL STANDARD STOCK SOLUTION - Accurately weigh
                    approximately 20.1 mg of 13C4-tetramine in a weigh boat and then transfer
                    into a 200 mL volumetric flask. Add 100 mL of acetonitrile and mix well
                    until  dissolved. Dilute to the 200 mL mark with additional acetonitrile
                    and mix well. The stock solution is stable for at least one year when
                    stored at -20 ±5°C.

             7.3.2.2 INTERNAL STANDARD PRIMARY DILUTION STANDARD (ISTD)
                    (500 ng/mL)  - Combine 50 jiL of the internal standard stock solution with
                    9.95 mL of deionized water in a 15 mL polypropylene centrifuge tube
                    (BD, Franklin Lakes, NJ) or equivalent.  The stock solution is stable for at
                    least one year when stored at -20±5°C.
                                          10

-------
7.3.3   ANALYTE STOCK STANDARD SOLUTIONS. Prepare or purchase three
       stock solutions using a reliable source of tetramine (Cambridge Isotopes, 50
       Frontage Road, Andover, MA 01810. CAS #:80-12-6; unlabeled material product
       #ULM-8147 and labeled material product # CLM-8146 (13C4 label)). These
       stock solutions are stable for at least one year when stored at -20±5°C.

       7.3.3.1 ANALYTE STOCK STANDARD SOLUTION 1 (SSI, 82.4 mg/L) -
             Accurately weigh approximately 20.6 mg of tetramine to a weigh boat and
             then into a 250 mL volumetric flask.  Add 100 mL of acetonitrile and mix
             well until dissolved. Dilute to the 250 mL with additional acetonitrile and
             mix well.

       7.3.3.2 ANALYTE STOCK STANDARD SOLUTION 2 (SS2, 206 ug/L) -
             Accurately transfer 25 uL of SSI into a 10 mL volumetric flask. Dilute
             with acetonitrile to 10 mL mark and mix well.

       7.3.3.3 ANALYTE STOCK STANDARD SOLUTION 3 (SS3, 8.24 mg/L) --
             Accurately transfer 1 mL of SSI into a 10 mL volumetric flask.  Dilute
             with acetonitrile to the 10 mL mark and mix well.

7.3.4   CALIBRATION STANDARD STOCK SOLUTIONS - Prepare the calibration
       standard stock solutions from dilutions of the analyte stock solutions in reagent
       water containing any preservatives required by site-specific circumstances (See
       Sects 2.2 and 8.1.3). For this purpose, a Falcon polypropylene 50 mL centrifuge
       tube (BD, Franklin Lakes, NJ) may be used by quantitatively transferring the
       volumes of the respective solution listed in the table below to the tube, diluting to
       the 40 mL mark, and mixing well.  (Note:  Diluting to the 40 mL mark provided
       sufficient accuracy in the developer's lab.  Other labs may wish to utilize alternate
       polypropylene vessels if they experience dilution related inaccuracies.)  The
       calibration curve is composed of at least six concentrations. These calibration
       standard solutions are stable for at least one year when stored at -20±5°C.

       7.3.4.1 PREPARATION OF CALIBRATION STANDARD STOCK
             SOLUTIONS - Calibrations standard stock solutions may be prepared
             using the volumes listed in Table 7-1 below.  The concentrations, along
             with the numbers of solutions, are  for illustration purposes only. Other
             concentrations may be required in  practice to meet performance and QC
             goal. (See Sect. 10.3 for the number of calibration solutions required for
             calibration.)
                                   11

-------
Table 7-1. Calibration Standard Stock Solution Volumes
Concentration
(ng/mL)
Total
Volume
(mL)
Stock Stock „, . „ . .
„ . „ . . Stock Solution
Solution 3 Solution 2 i r T ^
(uL) (uL)
0
0.5
1
2
5
10
15
25
50
75
100
250
40
40
40
40
40
40
40
40
40
40
40
97
194
388
24
49
73
121
24
36
49
121
7.3.5   QUALITY CONTROL SOLUTIONS - There are several types of quality control
       solutions, some of which are identical in composition but serve different QC
       functions and hence may be referred to by different names in Section 9.

       7.3.5.1   SECOND SOURCE QUALITY CONTROL SAMPLE - These samples
               are used to verify the accuracy of the calibration standard solutions
               (7.3.4) and are prepared the same way as the calibration standards.  They
               are prepared from an analyte source different than the calibration
               standard solutions as described more completely in Section 9.3.7.

       7.3.5.2   LABORATORY FORTIFIED BLANKS (LFBs) - LFBs are used
               throughout this method for various purposes.  The LFB is analyzed
               exactly like a sample, and its purpose is to verify that the methodology is
               be competently replicated, and that the laboratory has the capability to
               make accurate and precise measurements. The two specific LFBs are
               required in this method are referred to as LFB-low and LFB-high, which
               relate to initial and ongoing QC. For the demonstration of the method in
               the developer's laboratory, the LFB-low and -high are 5 and 75 ng/mL,
               respectively, prepared as indicated in Table 7.1, in Section 7.3.4. In a
               particular lab, the LFBs should be selected from similar points in their
               calibration range (e.g., LFB-low should be around 10 times the MRL
               (Sect. 9.2.4) and LFB-high should be around 150 times the MRL.

               The LFBs are inherently calibration standards and can be used to
               construct the calibration curve. However, the LFBs are  specifically used
                                   12

-------
                      to develop QC criteria during the initial demonstration of capability
                      (Sect. 9.2), and serve as an additional QC function during each analysis
                      batch. The LFBs serve a similar, but generally more stringent, QC
                      function as continuous calibration checks (Sect. 10.3).

             7.3.5.3   LABORATORY REAGENT BLANK. This blank is prepared as a LFB
                      with no analyte added (i.e., the 0 ng/mL in Table 7-1).
8.      SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1     SAMPLE VESSEL PREPARATION COLLECTION

       8.1.1   Samples can be collected in a 50-mL polypropylene vessel fitted with a flat-top
              polyethylene screw-cap (e.g., BD Falcon 50 mL centrifuge tube or equivalent).

       8.1.2   The performance data for the method presented in Section 13 are presented
              without addition of preservatives.  This is based on the stability of tetramine in the
              presence of preservatives suggested in Table 8-1, which suggests the analyte does
              not require sample preservation up to 28 days, particularly if tetramine is the sole
              analyte of interest in the sample.

       8.1.3   However, vessels should be prepared before sample collection with appropriate
              preservative(s) (Table 8-2) required by site-specific circumstances, e.g., to fulfill
              the purpose(s) listed in the Table 8-2. Preservation through binding free chlorine
              or dechlorination may also be necessary if analytical artifacts are observed in the
              samples but not the LFBs. All  initial and on-going QC requirements should be
              demonstrated for the preservatives added to the sample, particularly if added in
              combination. If tetramine is the only analyte determined, necessity of
              preservatives is expected to be  a very rare event.
                                           13

-------
Table 8-1. Recoveries of Tetramine in Preservatives over Time (n=3)
Water Type
Deionized
Chlorine
Monochloramine
Chlorine
Chlorine
Monochloramine
Chlorine
Chlorine
Chlorine
Deionized Water
Deionized Water
Preservative
-
-
-
Ammonium Chloride (0.1 g/L)
Sodium Thiosulfate (0.08 g/L)
Sodium Thiosulfate (0.08 g/L)
Sodium Sulfite (0.05 g/L)
Ascorbic Acid (0.1 g/L)
Ammonium Acetate (1.5 g/L)
Citric Acid (9.3 g/L)
Diazolidinyl Urea (1 g/L)
DayO
4°C
109 ±3
104 ±7
108 ±7
106 ±7
107 ±7
109 ±9
112±17
111±8
101 ±8
109 ±8
100 ±26
25°C
102 ±6
105 ±5
101±9
106 ±5
99 ±3
103 ±4
105 ±5
103 ±6
99 ±0
108 ±6
96 ±5
Day?
4°C
105 ± 14
107 ±9
100 ±2
104 ±5
96 ±2
99 ±6
110±6
101±7
116±6
104 ±1
102 ±6
25°C
112±4
111±5
108 ±4
113±8
111±5
111±5
105 ±5
111±10
111±3
110±5
103 ±2
Day 14
4°C
101±2
102 ± 1
101±4
116±4
100 ±2
113±2
101±2
104 ±3
114±2
95 ±2
99 ± 1
25°C
103 ±2
105 ±2
104 ±1
103 ±1
96 ±2
102 ±2
105 ±1
114±3
106 ±3
112±4
98 ± 10
Day 28
4°C
107 ±4
113±4
110±4
124 ±6
111±4
115±2
111±4
114± 11
107 ±2
114±4
109 ±8
25°C
108 ±4
118±2
112±10
112±5
115±9
113±4
109 ±10
111±4
110±3
110±2
110±6
                                                        14

-------
Table 8-2. Preservative Concentrations and Purposes of Preservatives
Compound
Ammonium
chloride
Ammonium
acetate
Sodium
thiosulfate
Sodium sulfite
Ascorbic acid
Citric acid
Diazolidinyl
urea
Mass added to
sample (mg)
5
75
4
2.5
5
465
50
Concentration in
sample (g/L)
0.1
1.5
0.08
0.05
0.1
9.3
1
Purpose
Binds free chlorine
Binds free chlorine
Dechlorinates free
chlorine and chloramine
Dechlorinates free
chlorine and chloramine
Dechlorinates free
chlorine and chloramine
pH adjustment
Microbial inhibitor
8.2    SAMPLE COLLECTION - When sampling from a water tap, samplers should
       request guidance about how long to flush the tap, if at all. Depending on site
       specific goals, incident managers may request that the tap not be flushed to
       minimize loss of contaminant.  If incident managers do not specify a shorter time,
       flush until the water temperature has stabilized (approximately 3-5 minutes).
       Collect samples from the flowing stream. It may be convenient to collect a bulk
       sample in a polypropylene vessel from which to generate individual 50 mL
       samples. Keep samples sealed from collection time until analysis. When
       sampling from an open body of water, fill the sample container with water from a
       representative area. Sampling equipment, including automatic samplers, should be
       free of plastic tubing, gaskets, and other parts that may leach interfering analytes
       into the water sample.

8.3    SAMPLE SHIPMENT AND STORAGE - Results of the sample storage stability
       study (Table 8-1) suggest that storage at 25°C produces results similar to reduced
       temperatures. As a matter of practice to ensure that the samples do not experience
       excessive temperature outside the stability range investigated, it is recommended
       that all samples be iced, frozen (-20°C), or refrigerated (4°C) from the time of
       collection until extraction. During method development, no significant differences
       were observed between standards that were frozen or refrigerated.

8.4    SAMPLE AND EXTRACT HOLDING TIMES - Results of the sample storage
       stability study (Table 8-1, n=3) suggest that tetramine has adequate stability for at
       least 28 days when collected, preserved, shipped, and stored as described in
       Sections 8.1, 8.2, and 8.3. As matter of practice, water samples should be
                                       15

-------
       extracted as soon as possible but must be extracted within 28 days. Data generated
       during this study indicates that extracts are stable for at least 28 days when stored
       at 0 °C or lower. As matter of practice, analysis should occur as soon as possible.
9.      QUALITY CONTROL

9.1     QC requirements include the initial demonstration of capability (IDC) and
       ongoing QC requirements that must be met when preparing and analyzing field
       samples. This section describes the QC parameters, their required frequencies,
       and the performance criteria that must be met in order to meet typical EPA quality
       objectives for drinking water analysis, although these objectives will be site
       specific during a remediation activity. These QC requirements are considered the
       minimum acceptable QC criteria in particular for this method which utilizes an
       isotopically labeled internal standard. Laboratories are encouraged to institute
       additional QC practices to meet specific needs [7].

9.2     INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be
       successfully performed prior to analyzing any field samples. Prior to conducting
       the IDC, the analyst must first generate an acceptable initial calibration following
       the procedure outlined in Section 10.2. It should be noted that the IDC is
       lengthier than some drinking water methods, but based on experience in the
       developer's lab, the IDC helps ensure successful long-term implementation of the
       method in a variety of other labs. Due to site-specific conditions during a
       environmental remediation activity, a shorter IDC may be necessary and
       appropriate. For example, a more minimal IDC could consist of: a)
       demonstration of low system background  (Sect. 9.2.1); b) 4-7 same-day replicates
       fortified near the midrange of the initial calibration curve for precision and
       accuracy demonstration, combined with c) the MRL estimation described in
       Section 9.2.4. However,  QC acceptance requirements, both initial (Sect. 9.2.1-
       9.2.4) and ongoing (Sect. 9.3) should not be changed,  and a shorter IDC may
       result in higher QC failure rates and less accurate quantitation in some
       concentration ranges. Labs should consider these risks before choosing a shorter
       IDC.

       9.2.1  INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND -
             Any time a new lot of solvents, reagents, and autosampler vials/plates are
             used, it must be demonstrated that an LRB is reasonably free of
             contamination and that the criteria in Section 9.3.1  are met.

       9.2.2  INITIAL DEMONSTRATION OF PRECISION - Prepare and analyze at
             least twenty replicates of both laboratory fortified blanks (LFB-high and
             LFB-low, see Sect. 7.3.5.2) over the course of at least 10 days.  Any
             sample preservative, as described in Section 8.1.2, must be added to these
             samples. For the initial demonstration of precision, the coefficient of
                                       16

-------
       variation for the concentrations of the replicate analyses must be less than
       20%.

9.2.3   INITIAL DEMONSTRATION OF ACCURACY - Using the same set of
       replicate data generated for Section 9.2.2, calculate the mean recovery. For
       the initial demonstration of accuracy, the mean recovery of the replicate
       values must be within ± 30% of the true value.

9.2.4   MINIMUM REPORTING LEVEL (MRL) ESTIMATION - Because
       clean-up goals will be site specific, laboratories need to estimate a
       minimum reporting level  so that incident managers can understand a
       specific laboratory's capabilities and can distribute samples to appropriate
       laboratories. Establishing the MRL concentration too low may cause
       repeated failure of ongoing QC requirements. If the IDC procedure (Sect.
       9.2.1-9.2.3) is followed explicitly, establishing the MRL as the lowest
       standard is expected to ensure compliance with QC requirements, This is a
       result of the rigor of the QC requirements in the lengthy IDC (Sect. 9.2.1-
       9.2.3), especially those associated with the LFBs (see Sect. 10.3.3).  If a
       shorter IDC is required by site specific conditions (see Sect. 2.2), the MRL
       should be confirmed with the procedure below.

        9.2.4.1  Fortify and analyze seven replicate LFBs at the proposed MRL
               concentration. These LFBs must contain all method
               preservatives described in Section 8.1.2. Calculate the mean
               measured concentration (Mean) and standard deviation for the
               method analytes in these replicates. Determine the half range for
               the prediction interval of results (HRpiR) for each analyte using
               the equation below:

                                  HRPIR = 3.9635

               where s is the standard deviation and 3.963 is a constant value
               for seven replicates.

        9.2.4.2  Confirm that the upper and lower limits for the prediction
               interval of result (PIR = Mean ± HRPIR) meet the upper and
               lower recovery limits as shown below:

            The Upper PIR Limit must be <150% recovery.

                                  HRP1R
                       FortifiedConcentration

            The Lower PIR Limit must be > 50% recovery.
                                 17

-------
                              FortifiedConcentration
                                                     lo0o/0>
              9.2.4.3  The MRL is validated if both the upper and lower PIR limits
                      meet the criteria described above (Sect. 9.2.4.2). If these criteria
                      are not met, the MRL has been set too low and must be
                      confirmed again at a higher concentration.
       9.2.5  CALIBRATION CONFIRMATION - The calibration is confirmed by
             analysis of a second source quality control sample as described in Section
             9.3.5
       9.2.6   DETECTION LIMIT (DL). This is a statistical determination of precision
             and accurate quantitation is not expected at this level.Replicate analyses
             for this procedure should be done over at least three days (i.e., both the
             sample preparation and the LC/MS/MS analyses should be done over at
             least three days).  At least seven replicate LFBs should be analyzed during
             this time period. The concentration may be estimated by selecting a
             concentration at two to five times the noise level.). The appropriate
             fortification concentrations will be dependent upon the sensitivity of the
             GC/MS system used. Any  preservation reagents added in Section 8.1.2
             must also be added to these samples. Note that the concentration for some
             IDC steps may be appropriate for DL determination, in which case the
             IDC data may be used to calculated the DL.  (For example, for the results
             presented in Section 13, twenty replicate LFBs were analyzed over 10
             days, e.g., three LFBs individually fortified on day one, two LFBs
             individually fortified on day two, and two LFBs individually fortified on
             day three, etc).  Analyze the replicates through all steps of Section 11.
             Calculate the DL from the  equation: DL = sxt(n.\)

             where:
                    s = standard  deviation of replicate analysis, without subtraction of
                    values of analyte free blanks
                    t = Student's t value for the 99% confidence level with n-l degrees
                           of freedom
                    n = number of replicates
9.3     ONGOING QC REQUIREMENTS - This section summarizes the ongoing QC
       criteria that must be followed when processing and analyzing field samples. The
       required QC samples for an analysis batch include the laboratory reagent blank
       (LRB) and four continuing calibration check (CCC) solutions.
                                       18

-------
9.3.1   LABORATORY REAGENT BLANK (LRB) - An LRB is required with
       each analysis batch (Sect. 3.1) to confirm that potential background
       contaminants are not interfering with the identification or quantitation of
       method analytes. Running the LRB first may prevent unnecessary
       analysis if the LRB is invalid. Preparation of the LRB is described in
       Section 7.3.5. If the LRB produces a peak within the retention time
       window of the analyte that would prevent the determination of the analyte,
       determine the source of contamination and eliminate the interference
       before processing samples. Background contamination must be reduced to
       an acceptable level before proceeding. Background from method analyte
       or other contaminants that interfere with the measurement of method
       analyte must be below 1/3 of the MRL. Blank contamination is estimated
       by extrapolation, if the concentration is below the lowest calibration
       standard.  This extrapolation procedure is not allowed for sample results as
       it may not meet data quality objectives. If the method analytes are detected
       in the LRB at concentrations equal to or greater than 1/3 the MRL, then all
       data for the problem analyte(s) must be considered invalid for all samples
       in the analysis batch.

9.3.2   ONGOING CALIBRATION. The analytical system in recalibrated at the
       beginning of each analysis batch using the  same analyte concentrations
       determined during the initial calibration. The acceptance criteria for the
       ongoing calibration is described in Section 10.2.5, except that removal of
       calibration points may result in too few calibration points and therefore  an
       invalid calibration.  The ongoing calibration is performed after the first
       two continuing calibration check (CCC) samples (Sec. 9.3.3) to  allow for
       corrective action if the calibration fails. As mentioned in Sect. 2.2, in
       some well considered circumstances and in consultation with the sample
       submitter about increased QC and quantitation risk, it may be desirable to
       not perform the ongoing calibration (Sect. 9.3.2) and instead rely on CCC
       samples (as described in Sect. 9.3.3) to verify ongoing calibration. If so,
       the beginning CCC of each analysis batch must be at or below the MRL in
       order to verify instrument sensitivity prior to any analyses. Subsequent
       CCCs should alternate between a medium and high concentration
       calibration standard.

9.3.3   CONTINUING CALIBRATION CHECK  (CCC) - CCC standards,
       containing the preservatives, if any, are analyzed at the beginning of each
       analysis batch, after every 20 field samples. Note that there are  up to four
       CCCs depending on the IDC appropriate for the site specific circumstance.
       In the lengthier IDC described in Sect. 9.2, there are four CCCs:  LFB-low
       and LFB-high, which are analyzed before the batch, and the lowest and
       highest calibration standards from the ongoing calibration (Sect 9.3.2),
       which are analyzed after the field samples. If this IDC approach is not
       appropriate, then there are at most two CCC standards, i.e. the calibration
       standards. Depending on site specific goals and tolerance of QC and
                                 19

-------
       quantitation risk, it may acceptable to only run one of these calibration
       standards as the CCC before and after the batch.  If so, the beginning CCC
       of each analysis batch must be at or below the MRL in order to verify
       instrument sensitivity prior to any  analyses.  Subsequent CCCs should
       alternate between a medium and high concentration calibration standard.
       See Section 10.3 for acceptance criteria for the various CCCs. Preparation
       of the CCCs is described in Section 7.3.5.

9.3.4   LABORATORY FORTIFIED BLANK (LFB) - Since this method utilizes
       procedural calibration standards, which are fortified reagent waters, there
       is no difference between the LFB and the CCC, except for the order in
       which they are run as part of an analysis batch and the corresponding QC
       acceptance criteria. The acronym LFB is  used for clarity in the IDC.

9.3.5   SECOND SOURCE QUALITY CONTROL SAMPLES (QCS) - As part
       of the IDC (Sect. 9.2), each time a new analyte stock standard solution 1
       (SSI, Sect. 7.3.3.1) is prepared, and at least quarterly, analyze a QCS
       sample from a source different from the source of the calibration
       standards. If a second vendor is not available, then a different lot of the
       standard should be used. The QCS should be prepared near the midpoint
       of the calibration range and analyzed as a CCC. Acceptance criteria for the
       QCS are identical to the CCCs; the calculated amount for each analyte
       must be ± 30% of the expected value. If measured analyte concentrations
       are not  of acceptable accuracy, check the entire analytical procedure to
       locate and correct the problem.

9.3.6   INTERNAL STANDARD (IS) - The analyst must monitor the peak area
       of the IS in all injections during each analysis day. The IS peak area  must
       meet the criteria in the both following two subsection

       9.3.6.1   The internal standard should produce a peak area at least five
               times higher than the peak area  of the quantitation ion of
               tetramine in the lowest concentration calibration solution. If it
               does not, the concentration of IS may not be as predicted.
               Prepare new calibrations solutions, QC samples, and field
               samples with an appropriately increased concentration of IS.

       9.3.6.2   The IS response (peak area) in any chromatographic run must not
               deviate from the response in the most recent CCC by more than
               30%, and must not deviate by more than 50% from the area
               measured during initial analyte  calibration. If the IS area in a
               chromatographic run does not meet these criteria, inject  a second
               aliquot of that extract.

       9.3.6.2.1 If the reinjected aliquot produces an acceptable IS response,
               report results for that aliquot.
                                20

-------
       9.3.6.2.2 If the reinjected aliquot fails the IS criterion, the analyst should
               check the calibration by reanalyzing the most recently acceptable
               calibration standard. If the calibration standard fails the criteria
               of Section 10.3, recalibration is in order per Section 10.2. If the
               calibration standard is acceptable, report results obtained from
               the reinjected aliquot, but annotate as "suspect/IS recovery."
               Alternatively, prepare another aliquot of the sample as specified
               in Section 11.2 or collect a new sample and re-analyze.
9.3.7   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) and LFSM
       DUPLICATES - The isotopically labeled internal standard in this method
       also serves the role of the LFSM, which is used to determine that the
       sample matrix does not adversely affect method accuracy. In the context
       of application of this method for environmental restoration, it is not
       expected that there would be a native tetramine background concentration.
       Also, it is likely that the water samples will come from the same drinking
       water system, and  hence the sample matrices from a single collection time
       will be very similar.  Further, experience with the automated extraction
       equipment used suggests that if most failures in IS QC requirements result
       from failure of the automation equipment. This would correspond to
       LFSM failure, as well. Accordingly, neither LFSMs or duplicate LFSMs
       would be expected to yield additional information about influence of
       sample matrix on method accuracy, except for the unlikely case of a
       feature of the sampling/restoration plan that produces a co-eluting peak
       with identical chromatographic and mass spectral properties as tetramine.
       In this case, the lab should discuss with the submitter.

       9.3.7.1    If an LFSM and LFSM is deemed necessary, calculate the
                relative percent difference (RPD) for duplicate LFSMs (LFSM
                and LFSMD) using the equation
                                 \LFSM -LFSMD\
                             = -±	^
                               (LFSM + LFSMD} I'2
       9.3.7.2   Relative percent difference (RPD) for duplicate LFSMs should
                be <30% for samples fortified at or above their native
                concentration.  Greater variability may be observed when
                LFSMs are fortified at analyte concentrations that are within a
                factor of two of the MRL. LFSMs fortified at these
                concentrations should have RPDs that are <50%.  If the RPD of
                any analyte falls outside the designated range, and the
                laboratory performance for that analyte is shown to be in control

                                21

-------
                             in the CCC, the recovery is judged to be matrix biased.  The
                             result for that analyte in the unfortified sample is labeled
                             "suspect/matrix" to inform the data user that the results are
                             suspect due to matrix effects.

             9.3.5  FIELD DUPLICATE (FD) - Field duplicates check the precision
                   associated with sample collection, preservation, storage, and laboratory
                   procedures.  Some of these factors are out of control of the laboratory, and
                   the rest are covered by other QC checks.  Accordingly, results of any field
                   duplicates requested  should be discussed with the sample submitter if they
                   do not meet the following criteria.

                    9.3.5.1 Calculate the relative  percent difference (RPD) for duplicate
                          samples (FD1 and FD2) using the equation

                                                      FDI-FD2
                                             RPD=-,	r—xlOO
                                                          FD2}I2

                    9.3.5.2 RPDs for FDs should be <30%. Greater variability may be
                           observed when FDs have analyte concentrations that are within a
                           factor of two of the MRL. At these concentrations, FDs should
                           have RPDs that are <50%.  If the RPD of any analyte falls outside
                           the designated range, and the laboratory performance for that
                           analyte is shown to be in control in the CCC, the recovery is
                           judged to be biased.  The result for that analyte in the unfortified
                           sample is labeled "suspect/field duplicate bias" to inform the data
                           user that the results are suspect due to field bias. (Note some other
                           sources of lab bias may also be present.)
10.     CALIBRATION AND STANDARDIZATION

       10.1   All laboratory equipment should be calibrated according to manufacturer's
             protocols and equipment with expired calibrations should not be used.
             Demonstration and documentation of acceptable mass spectrometer tune and
             initial calibration is required before any samples are analyzed. After the initial
             calibration is successful, the instrument is recalibrated using the same conditions
             as the initial calibration before each analysis batch. After the batch, the lowest
             and highest calibration solutions are run as continuing calibration checks (CCC)
             Verification of mass spectrometer tune must be repeated each time a major
             instrument modification is made or maintenance is performed, and prior to analyte
             calibration.

       10.2   INITIAL CALIBRATION
                                              22

-------
10.2.1  MS TUNE - Calibrate the mass and abundance scales of the MS
       with calibration compounds and procedures prescribed by the
       manufacturer with any modifications necessary to meet tuning
       requirements. For an Agilent MSD, some labs have experienced
       better results if following the automatic tune, they perform a
       manual tune to set the mass resolution to unit mass, the peak width
       to 0.50 ± 0.01 amu and the abundance for the ion at mass 69 to
       500,000 ± 50,000 counts. For other instruments, follow
       manufacturer's protocols to tune the instrument.

10.2.2  INSTRUMENT CONDITIONS - Operational conditions are
       tabulated in Section 6.8.3.  Alteration of the conditions is not
       recommended and would require redevelopment of QC criteria.
       Frequently reported problems can be avoided by: 1) checking that
       needle wash solutions are adequately filled and the injection
       syringe is functioning properly and 2) changing the septum and
       inlet liner as needed.

10.2.3  Prepare six calibration standards as described in Section 7.3.4.
       Note that as procedural calibration standards, they are processed
       through the procedure in Section 11, in which the isotopically
       labeled internal standard is added before extraction.  In practice,
       the lowest concentration of the calibration standard must be at or
       below the MRL (Sect.  9.2.4), which will depend on system
       sensitivity. The lowest point on the calibration curve is close to the
       reported detection limit and the highest point is above the expected
       range of results.  The remainder of the points are distributed
       between these two extremes, with the majority of points in the
       concentration range where most unknowns are expected to fall.

10.2.4  The GC/MS system is  calibrated using the internal standard
       technique, as implemented by the  data system software. Construct
       a calibration curve using at least a  six-point curve of response
       ratios (i.e., ratio of calibration standard peak area to internal
       standard peak area).  As the internal standard concentration is
       consistent among samples and calibrators, some labs have found it
       convenient to set it to a value of one instead of the actual
       concentration.

10.2.5  CALIBRATION ACCEPTANCE -- Calculate the slope and
       intercept of the calibration curve with 1/x weighting (or other
       appropriate weighting) by a linear least squares fit (or other
       appropriate calibration function).  Evaluate the r2 value for the
       curve, which must be greater than 0.990.  Linearity  of the standard
       curve should extend over the entire standard range.  The intercept
       should not be significantly  different from 0; if it is, the source of
                          23

-------
                    the bias should be identified.  Each calibration point, except the
                    lowest point, for the analyte should calculate to be 70 to 130 percent
                    of its true value. The lowest point should calculate to be 50 to 150
                    percent of its true value. If these criteria cannot be met, the analyst
                    will have difficulty meeting ongoing QC criteria.  If any standard
                    is in error and does not fit the standard curve (i.e., the r  value for
                    the curve is < 0.990), it can be removed from the calibration. No
                    more than one standard may be discarded in any given calibration
                    curve.  If either the high or low standard is dropped, the reporting
                    limits must be adjusted accordingly.  The resulting r2 value must be
                    greater than 0.990.

10.3   CONTINUING CALIBRATION CHECKS (CCCs). As described in Sect 9.3.3,
       up to four CCCs are used in conjunction with each analysis batch. If applicable,
       LFB-low and LFB-high are run at the beginning of the batch, and the calibration
       solutions are run at the end. The LFBs serve to verify the initial IDC, and the
       calibration solutions verify the calibration generated at the start of the analysis.
       The LRBs, LFBs, and CCCs are not counted as the 20 samples that constitute an
       analysis batch.

              10.3.1 Inject an aliquot of the appropriate concentration calibration
                    solution  and analyze with the same conditions used during the
                    initial calibration.

              10.3.2 Acceptance of the calibration solutions is based on the same
                    criteria as described in Section 10.2.5. Failure to meet these
                    criteria is a rare occurrence, and suggests maintenance of the
                    GC/MS system is required.

              10.3.3 Acceptance of the results of the LFB-Low and LFB-High is based
                    on the Quality Control Limits (Sect.  10.3.3.1) established via the
                    IDC. Acceptability of results for that entire analytical batch is
                    dependent upon the agreement of the results from these control
                    materials within established ranges. Quality Control Limits for the
                    CCCs  are based primarily on the standard deviation (on-i, sigma)
                    of the replicate analysis in the IDC (Sect. 9.2.2).  Section 13.3
                    presents  sample values for these parameters obtained in the
                    developer's laboratory, in which 20 replicate analyses performed
                    over no less than 10 days are used to establish the LFB-low and -
                    high limits (Sect. 9.2.2). If the CCC results do not meet the
                    following criteria, it is "out-of-control," and the cause of the failure
                    must be determined and corrected.  No results from the associated
                    analytical batch may be reported. These criteria apply to non-zero
                    analyte concentrations used to make the quality control  solutions in
                    section 7.3.5.1
                                        24

-------
                    10.3.3.1   If both of the LFB-Low and LFB-High results are
                              within 2on-i of the mean determined during the IDC,
                              then accept the entire analytical batch. Otherwise,
                              reject the entire analytical batch.
              10.3.4 Common remedial actions if the CCCs fails to meet acceptable
                    criteria

              10.3.4.1   LOW ANALYTE RESPONSE - If the signal-to-noise of the
                       low standard confirmation ion falls below 10, this indicates that
                       the instrumental sensitivity, or SPE recovery, has fallen below
                       acceptable limits.  The following steps should be taken and the
                       instrument sensitivity rechecked after each corrective action is
                       performed.  Once sensitivity has been reestablished, further
                       steps are not necessary.

                       i.   Re-extract the samples.
                       ii.   If tailing is a significant issue, clip the GC column.
                       iii.  Ensure the filament for the MS is still intact.
                       iv.  Clean the mass spectrometer source
                       v.   Clean the gas chromatograph inlet liner

              10.3.4.2   Analyte in standards - If an inordinately large amount of
                       analyte is measured in one of the calibration standards, but this
                       is not seen in the remainder of the samples, this indicates a
                       contamination of this particular sample. The source of this
                       incident should be investigated to prevent repeat occurrences,
                       but no further action is required. The contaminated calibration
                       standard should be excluded when developing the calibration
                       curve.

              10.3.4.3   Analyte in all samples - If an inordinately large amount of
                       analyte is present in all measurements for a particular day, it is
                       likely that one or more of the spiking solutions are
                       contaminated. If necessary, prepare new solutions.

11.    PROCEDURE

            11.1    SAMPLE PREPARATION

              11.1.1 Samples are preserved, collected and stored as presented in Section
                    8. Allow samples to come to room temperature prior to analysis.

              11.1.2 If using a TurboVap 96 evaporator system, set it to 65-75°C.
                    Follow manufacturer's direction for other equipment.
                                       25

-------
Note:  Steps 11.1.3 through 11.1.13 can be performed using an automated
liquid handler or a manual pipettor with a manual 96-well manifold.
However, data presented in this document was collected using an
automated liquid handler.

11.1.3  Fill 96 plate wells.
       11.1.3.1  Into each well of the 96-well Nunc deep well plate
                (Nalge Nunc International, Rochester, NY), add 50 uL
                of the isotopically-labeled internal standard (refer to
                section 7.3.2)
       11.1.3.2  Into each sample well, add 1000 uL of sample.
       11.1.3.3  Into each blank well, add 1000 uL of reagent water (for
                the LRB).
       11.1.3.4  Into each calibration standard well, add 1000 uL of
                tetramine calibration standard stock solutions (refer to
                sections 7.3.4)
       11.1.3.5   Into each quality control well, add 1000 uL of
                appropriate quality control material,  (refer to section
                7.3.5)

11.1.4  Mix on the plate shaker for 2 min or by other appropriate means.

11.1.5  Plate SPE procedure. For each well on the Nunc plate filled in
       Section 11.1.3, perform the following steps and do not let wells go
       dry for more than  1 minute:

       11.1.5.1 Condition/preclean the selected well on the
               Phenomenex® Strata-X 60-mg SPE well plate
               (Phenomenex, Torrance, CA) with 1125 uL of 100%
               methanol.

       11.1.5.2 Condition the SPE plate with 1125 uL of deionized
               water.

       11.1.5.3 Load 1000 uL from the Nunc plate and draw through the
               SPE plate using positive or negative pressure.

       11.1.5.4 Wash the SPE plate with 1125 uL of 5% methanol/95%
               water (Sect. 7.2.1).

       11.1.5.5  Elute the sample with 800 uL  of acetonitrile into a 96-
               well Nunc deep well plate.

11.1.6  Blow down the sample to dryness using nitrogen gas at 65-75°C.
       If using a TurboVap evaporator system, set the flow rate to 45 flow
                          26

-------
        units until approximately 50% has been evaporated. Then raise the
        flow rate to 75 flow units until dry. When using systems other than
        the TurboVap, set the flow rate for the blow down gas according to
        manufacturer's directions.

  11.1.7  Add 100 uL of acetonitrile to reconstitute each sample and vortex.

  11.1.8   Transfer the acetonitrile solution into appropriate autosampler
          vials or a 96-well autosampler plate.

11.2   ANALYSIS OF SAMPLE EXTRACTS

  11.2.1 Establish operating conditions as described in Section 10.2.2.

  11.2.2 Establish a valid initial calibration following the procedures
        outlined in Section 10.2 or confirm that the calibration is still valid
        by running both CCCs as described in Section 10.3. If establishing
        an initial calibration for the first time, complete the IDC as
        described in Section 9.2.

  11.2.3 Set up the available automation equipment and software as
        specified by the manufacturer for batch analysis, paying particular
        attention to the following frequent stumbling blocks:

        11.2.3.1 On the instrument computer, edit the automation
                 software:
               (a)  Select the sample type.
               (b) Identify the correct vial position.
               (c) Name the sample.  Due to large number of samples
               analyzable with the automation equipment, it is important
               that appropriate record keeping (e.g., database, notebooks,
               data files, etc.) should be used to track specimens.
               (d) Enter information related to particular specimens into
               the software manually or by electronic transfer.
               (e)  Select the instrument control method.
               (f) Identify the target path where the data will be stored.

         11.2.3.2 Check to be sure that the number and positions of
                 samples entered on the sequence set-up page correspond
                to the samples in the autosampler.

  11.2.4  Run the automation sequence to analyze the batch of aliquots of
        field and QC samples at appropriate frequencies (Sect. 9, 10.3).
        All field, QC, and calibration standards should be run using the
        same GC/MS conditions. At the conclusion of data acquisition, use
        the same software that was used in the calibration procedure to
                           27

-------
                    identify the peaks in predetermined retention time windows of
                    interest. Use the data system software to examine the ion
                    abundances of components of the chromatogram.

              11.2.5 COMPOUND IDENTIFICATION - The presumed tetramine peak
                    in the sample must appear in the same retention time window as
                    the isotopically-labeled internal standard (around 11.6 min in the
                    developer's lab) and have similar chromatographic characteristics
                    such as peak shape. This relies on expert judgment of the analyst
                    since the retention times reported by the software are not always
                    reliable.  Identification of the peak as tetramine is then confirmed
                    through calculating the confirmation ratio (CR), i.e., by dividing
                    the response for m/z 240 by the response for m/z 212 of the
                    presumed tetramine peak. Using the manufacturer's software or
                    manually, compare the confirmation ratio of the peak from the
                    sample with the mean of the CRs measured for the six calibration
                    standards associated with that batch. The mean CR is the average
                    CR from the calibration standards only and is batch dependent.
                    The CR value for each sample should be within 30% of the mean.
                    (CR value was 1.74 in the developer's lab for the IDC samples).
12.     DATA ANALYSIS AND CALCULATIONS

            12.1   Concentrations are calculated using the ions listed in Section 6.8.3.
                  Use of other ions is not advised. If a particular instrument cannot
                  produce the fragments listed in section 6.8.3, this instrument should
                  not be used to run this method.

            12.2   Calculate analyte concentrations using the ongoing multipoint
                  calibration established in Section 9.3.2. Do not perform calibration
                  using just the CCC or LFB-low and -high data to quantitate analytes
                  in samples, although these samples might be part of the ongoing
                  calibration curve.

            12.3   All raw data files are quantified using the quantitation capabilities of
                  the instrument software.  The peaks are automatically integrated
                  using the software-associated integration program, and the
                  integration of each peak is reviewed and manually corrected as
                  appropriate. This is particularly important for the calibration
                  standards. The quality control samples (e.g., CCCs and LFBs) are
                  quantified and evaluated against the calibration curve, and each field
                  sample is then quantified against that calibration curve. The run data
                  can be processed within instrument data analysis software and
                  exported to external spreadsheets, per laboratory policy, generating
                                       28

-------
     files containing the unknown and QC concentrations, retention
     times, standard curves, and other run information.

12.3.1  Results are generally reported to two significant digits. In addition
       to analytical measurements of unknowns, statistical results of
       measurement of blanks should accompany all results.

12.3.2  Check all sample and analytical data for transcription errors and
       overall validity after being entered into the instrument software
       database. Back up onto external media both the instrument and
       data storage databases according to laboratory guidelines.
                          29

-------
13.   METHOD PERFORMANCE
            13.1   ANALYTICAL IDENTIFICATION-Analyte identification using
                  the approach described in Section 11.2.5 resulted in no false
                  positives or negatives for the samples reported below.  There was
                  very low background noise according to the signal-to-noise ratios for
                  the m/z monitored.

            13.2   SINGLE LABORATY MINIMUM REPORTING LEVELS and
                  DETECTION LIMIT- The reportable range of results for tetramine
                  is summarized below,  along with the DL determined from the IDC
                  procedure (n=20, >10 days) described previously. The lowest
                  standard is used as the method reportable limit, and the DL
                  calculated from the standard deviation of replicate measurements of
                  that standard (in the case of Table 13-1, 0.059). The highest
                  reportable limit is based on the highest linear standard.
Table 13-1. Method Performance
Compound
Tetramine
(retention time =
11.6 min)
Minimum reporting
level (ng/mL)
0.5
Highest reportable
limit (ng/mL)
250
Method DL (ng/mL)
0.15
            13.3   SINGLE LABORATORY ACCURACY AND PRECISION for
                  LFBs - Single lab precision and accuracy data is represented in
                  Table 13-2. Accuracy is defined as the mean of the measured
                  concentration in the fortified samples divided by the fortification
                  concentration, expressed as a percentage. Method accuracy was
                  determined by analyzing LFBs at the two non-zero levels in Section
                  7.3 (i.e., LBF-low and -high) and twenty analyses for each of the
                  two concentration levels were completed over a period of 28 days.
                  The means, standard deviations, and relative standard deviations for
                  the two LFBs are shown in Table 13-2. The means are less than one
                  standard deviation from the known concentration.

Table 13-2. Single Lab Precision and Accuracy Data
Analyte
Tetramine
Sample
LFB-low
LFB-
high
Fortified
Concentration
(Mg/L)
5
75
Mean of
IDC
Replicates
(Mg/L)
5.03
75.5
Standard
Deviation
(Mg/L)
0.28
2.8
RSD
(%)
5.6
3.7
Accuracy
of Mean
(%)
100
101
                                      30

-------
            13.4   SINGLE LABORATORY RECOVERY AND PRECISION FOR
                   TAP WATER MATRICES.  Table 13-3 expresses percent mean
                   recoveries for tetramine in several different chlorinated and
                   chloraminated tap waters derived from the types of sources (i.e.,
                   ground or surface water) indicated. Water quality parameters
                   describing these sources are indicated in the footnotes. Percent
                   recoveries were determined by dividing the measured concentration
                   by the spiked concentration (75 ug/L) (n=3 for each day). No
                   tetramine or interferences were detected in the unspiked samples.
            13.5   SAMPLE STORAGE STABILITY STUDIES - Table 13-3 also
                   presents tetramine storage stability data. Samples were collected and
                   stored as described in Section 8 and also at room temperature.  No
                   preservatives were added to the samples (See Sect. 8.1.2). The
                   precision and average recovery of triplicate analyses was conducted
                   on Days 0, 7, 14 and 28. These data support the maximum 28 day
                   aqueous holding time specified in Section 8.4.
Table 13-3. Percent Recovery of Tetramine for Several Tap Water Matrices and
Residual Disinfectants
Water Type
Ground Water la
(chlorine)
Surface Water 2b
(chloramine)
Surface Water 3C
(chlorine)
Surface Water 4d
(chloramine)
Surface Water 5e
(chlorine)
DayO
4°C
103 ±8
107 ±8
110±3
102 ±2
102 ±3
25°C
109
100
96
94
99
±3
±1
±0
±4
±7
Day?
4°C
105
99
102
102
117
±3
±3
±3
±1
±4
25°C
113
114
106
107
108
±4
±3
±6
±3
±3
Day 14
4°C
113±4
92 ±3
108 ±6
98 ±7
108 ±6
25°C
104 ±5
105 ±6
107 ±5
98 ±5
108 ±6
Day 28
4°C
122 ±2
119±6
111±2
109 ±2
112±3
25°C
114±7
111±3
107 ±2
114±3
116±8
aTotal organic carbon (TOC) not detected in well-field; pH 7.6; hardness 350 mg/L; Chlorine 0.2-0.4 mg/L;
(monthly averages)
 TOC 7.61 mg/L; pH 9.2; hardness 65 mg/L; Monochloramines 2.4 mg/L (monthly averages)
°TOC 2.0 mg/L; pH 7.3; hardness  135 mg/L; Chlorine 1 mg/L (monthly averages)
 TOC 2.3; pH 7.4; hardness 190 mg/L; Monochloramine 3.4 mg/L (monthly averages)
eTOC 1.0; pH 8.5; 130 mg/L; Chlorine 0.8 mg/L (monthly averages)
                                        31

-------
14.     POLLUTION PREVENTION

            14.1   This method utilizes solid phase extraction to extract analytes from
                  water. It requires the use of reduced volumes of organic solvent and
                  very small quantities of pure analytes, thereby minimizing the
                  potential hazards to both the analyst and the environment as
                  compared to the use of large volumes of organic solvents in
                  conventional liquid-liquid extractions.

            14.2   For information about pollution prevention that may be applicable to
                  laboratory operations,  consult "Less is Better: Laboratory Chemical
                  Management for Waste Reduction" available from the American
                  Chemical Society's Department of Government Relations and
                  Science Policy on-line at
                  http ://portal. acs. org/portal/fileFetch/C/WPCP_012290/pdf/WPCP_0
                  12290.pdf (accessed May 2010).

15.     WASTE MANAGEMENT

            15.1   Dispose of waste materials in compliance with the laboratory
                  chemical hygiene plan, as well  as federal, state, and local
                  regulations.  Always dispose of solvents and reagents in an
                  appropriate container clearly marked for waste products and
                  temporarily store them in a chemical fume hood.  Dispose of
                  tetramine in an  appropriate waste stream as well.  Tetramine is not
                  destroyed by autoclaving [8], so wash any other non-disposable
                  glassware, empty ampoules, and/or apparatus before recycling or
                  disposing of in an appropriate manner.
                                       32

-------
16.    REFERENCES

1.      Barrueto, F., Jr.; L.S.Nelson; R.S. Hoffman; M.B. Heller; P.M.  Furdyna; et al.
       Poisoning by an Illegally Imported Chinese Rodenticide Containing
       Tetramethylenedisulfotetramine. Morbidity and Mortality Weekly Report
       (MMWR). 52(10): p. 199-201.
2.      Whitlow, K.S., B. Belson, F. Buarueto, L. Nelson and A.K. Henderson.
       Tetramethylenedisulfotetramine: old agent and new terror.  Ann Em erg Med,
       2005. 45(6): p. 609-13.
3.      Chau, C.M., A.K. Leung, and I.K.  Tan, Tetraminepoisoning. Hong Kong Med J,
       2005. 11(6): p. 511-4.
4.      "Carcinogens - Working With Carcinogens," Department of Health, Education,
       and Welfare, Public Health Service, Center for Disease Control, National Institute
       for Occupational Safety and Health, Publication No. 77-206, August 1977.
5.      "OSHA Safety and Health Standards, General Industry," (29CFR1910),
       Occupational Safety and Health Administration, OSHA 2206, (Revised, January
       1976).
6.      "Safety in Academic Chemistry Laboratories," American Chemical Society
       Publication, Committee on Chemical Safety, 7th Edition. Information on
       obtaining a copy is available at http://membership.acs.org/C/CCS/pub_3.htm
       (accessed June, 2009).  Also available by request at OSS@acs.org.
7.      Taylor, J.K., Quality Assurance of Chemical Measurements. 1987, Boca Raton:
       Lewis Publishers.
8.      Li, X-h., Z-h Chen, Y-f Lu , X-w. GE, and J. Quo.Safety destruction of
       tetramethylene disulfotetramine and its medical waste. Chin J Prev Med, 2005.
       39(2): p. 88-90.
                                       33

-------
United States
Environmental Protection
Agency
PRESORTED STANDARD
 POSTAGE & FEES PAID
         EPA
   PERMIT NO. G-35
Office of Research and Development (8101R)
Washington, DC 20460

Official Business
Penalty for Private Use
$300

-------