United States      Prevention, Pesticides     EPA712-C-96-042
          Environmental Protection    and Toxic Substances     August 1996
          Agency        (7101)
&EPA    Product Properties
          Test Guidelines
          OPPTS 830.7860
          Water Solubility
          (Generator Column
          Method)

-------
                           INTRODUCTION
     This guideline is one  of a series  of test guidelines  that have been
developed by the Office of Prevention, Pesticides and Toxic Substances,
United States Environmental Protection Agency for use in the testing  of
pesticides and toxic substances, and the  development of test data that must
be submitted to the Agency  for review under Federal regulations.

     The Office of Prevention, Pesticides and Toxic Substances (OPPTS)
has  developed this guideline through  a  process of harmonization that
blended the testing  guidance and requirements that existed in the Office
of Pollution Prevention and Toxics  (OPPT) and appeared in Title 40,
Chapter I,  Subchapter R of the Code of Federal Regulations  (CFR), the
Office of Pesticide Programs (OPP) which appeared in publications of the
National Technical  Information Service (NTIS) and  the guidelines pub-
lished by the Organization  for Economic Cooperation and Development
(OECD).

     The purpose of harmonizing these guidelines into a single set  of
OPPTS  guidelines is to minimize variations among the testing procedures
that must be performed to meet the data  requirements of the U. S. Environ-
mental Protection Agency  under the Toxic  Substances Control Act (15
U.S.C. 2601) and the Federal Insecticide,  Fungicide and Rodenticide Act
(7U.S.C. I36,etseq.).

     Final  Guideline Release: This document is available from the U.S.
Government Printing Office, Washington, DC 20402 on The Federal Bul-
letin  Board.   By  modem  dial   202-512-1387,   telnet   and  ftp:
fedbbs.access.gpo.gov     (IP     162.140.64.19),    internet:     http://
fedbbs.access.gpo.gov, or call 202-512-0132 for disks or paper copies.
This guideline is available in ASCII and PDF (portable document format)
from the EPA Public Access Gopher (gopher.epa.gov) under the heading
"Environmental Test Methods and Guidelines."

-------
OPPTS 830.7860  Water solubility (generator column method).
     (a) Scope—(1) Applicability. This guideline is intended to meet test-
ing  requirements  of both  the  Federal  Insecticide,  Fungicide,  and
Rodenticide Act (FIFRA) (7 U.S.C. 136, et seq.) and the Toxic Substances
Control Act (TSCA) (15 U.S.C. 2601).

     (2) Background. The source materials used in developing this har-
monized OPPTS test guideline are the OPPT guideline  under 40 CFR
796.1860 Water solubility (generator column method), and OPP guideline
63-8 Solubility (Pesticide Assessment Guidelines, Subdivision D: Product
Chemistry, EPA Report 540/9-82-018,  October 1982).

     (b) Introduction—(1) Purpose, (i) The water solubility of a chemical
is  defined as the equilibrium  concentration of the chemical in a saturated
aqueous solution at a given temperature and pressure. The aqueous phase
solubility is an important  factor in governing the movement,  distribution,
and rate of degradation of chemicals in the  environment.  Substances that
are relatively water soluble are more  likely to be  widely distributed by
the hydrologic  cycle  than those which are  relatively insoluble. Further-
more, substances with higher water solubility  are more likely to undergo
microbial or chemical degradation in the environment because dissolution
makes them "available" to interact and, therefore, react with other chemi-
cals  and microorganisms.  Both the extent  and rate of degradation via hy-
drolysis, photolysis,  oxidation, reduction,  and biodegradation depend on
a chemical being soluble in water (i.e., homogeneous kinetics).

     (ii) Water  provides the medium in which many organisms live, and
water is a major component of the internal environment of all living orga-
nisms (except for  dormant stages  of certain life forms).  Even organisms
which are adapted to life in a gaseous  environment require water for nor-
mal  functioning. Water is thus the medium  through which  most  other
chemicals are transported to and into living cells.  As a result, the extent
to  which  chemicals  dissolve  in water will be a  major determinant for
movement through the environment and entry into living systems.

     (iii) The water solubility of a chemical also has an effect on its sorp-
tion  into and desorption from  soils and sediments, and on volatilization
from aqueous media.  The more soluble a chemical substance is, the less
likely it is to sorb to soils and sediments and the less likely it  is to volatil-
ize from water. Finally, the design of most chemical tests and many eco-
logical and health tests requires precise knowledge of the  water solubility
of the chemical to be tested.

     (2) Definitions and units,  (i) Concentration of a  solution is  the
amount of solute in a given  amount of solvent or solution  and can be
expressed  as a weight/weight or weight/volume relationship. The conver-
sion from a weight relationship to one of volume incorporates density as
a factor. For dilute aqueous solutions, the density of the solvent is approxi-

-------
mately equal to the density of the  solution; thus, concentrations expressed
in milligrams per liter (mg/L) are approximately equal to  10'3 g/103 g
or parts per  million  (ppm); those expressed  in  micrograms  per  liter
(|-ig/L) are  approximately equal to 10-6 g/103 g or parts per billion (ppb).
In addition, concentration can be expressed in terms of molarity, normality,
molality, and mole fraction. For example, to convert  from weight/volume
to molarity molecular mass is incorporated as a factor.

     (ii) Density is the  mass of a unit volume of a material. It is a function
of temperature, hence  the temperature at which it is  measured should be
specified. For a solid,  it is the density of the impermeable portion rather
than the bulk density. For solids and liquids, suitable units of measurement
are grams per cubic centimeter (g/cm3). The density of a solution is the
mass of a unit volume of the solution and suitable units of measurement
are grams per cubic centimeter.

     (iii) A saturated solution is a solution in which  the dissolved solute
is in equilibrium with an excess  of undissolved solute; or a solution in
equilibrium such that at a fixed temperature and pressure, the concentration
of the solute in the solution is at its maximum value  and will not change
even in the presence of an excess of solute.

     (iv) A solution is  a  homogeneous mixture  of two or more substances
constituting a single phase.

     (v) A generator column is used to produce or generate saturated solu-
tions of a solute  in a solvent. The column (see figure 1 under paragraph
(c)(l)(i)(A) of this guideline) is packed with a solid  support coated with
the solute, i.e., the organic compound whose solubility is to be determined.
When water  (the solvent) is  pumped through the column,  saturated solu-
tions of the solute are generated.  Preparation of the  generator column is
described under paragraph (c)(l)(i)(A) of this guideline.

     (vi) An  extractor  column is used to extract the solute from the satu-
rated solutions produced by the generator column. After extraction onto
a chromatographic support, the solute is eluted with a solvent/water mix-
ture  and subsequently  analyzed by high pressure liquid chromatography
(HPLC), gas  chromatography (GC), or any other suitable analytical proce-
dure. A  detailed description of the preparation of the extractor  column
is given in paragraph (c)(l)(i)(D) of this guideline.

     (vii) The sample loop is a Vie in O.D. (1.6 mm) stainless steel tube
with an internal volume between  20 and 50 (iL. The loop is attached to
the sample injection valve  of the HPLC and  is used to  inject  standard
solutions into the mobile phase of the HPLC when  determining the re-
sponse factor for the recording integrator.  The exact volume of the loop
must be determined as  described in paragraph (c)(3)(ii)(B)(7) of this guide-
line when the HPLC method is used.

-------
     (viii) The  response factor (RF) is the  solute concentration required
to give a 1 unit area chromatographic peak or 1 unit output from the HPLC
recording integrator at a particular recorder attenuation. The factor is re-
quired to convert from units of area to units of concentration. The  deter-
mination of the response factor is given  in paragraph (c)(3)(ii)(B)(2) of
this guideline.

     (3) Principle of the test method, (i) This test method is based on
the  dynamic coupled column liquid chromatographic (DCCLC) technique
for  determining the aqueous solubility of organic compounds that was ini-
tially developed by May et al. (see paragraphs (e)(5) and  (e)(6) of this
guideline), modified by DeVoe et al. (see paragraph (e)(l) of this guide-
line), and finalized by Wasik  et al. (see paragraph (e)(ll) of this guide-
line). The DCCLC technique utilizes  a generator column, extractor column
and HPLC coupled or interconnected to provide a continuous closed flow
system. Saturated  aqueous solutions of the test compound  are produced
by  pumping  water through the generator column that is packed with  a
solid support coated with the compound. The compound is extracted from
the  saturated solution onto an extractor column, then eluted from the ex-
tractor column with a solvent/water mixture and subsequently analyzed
by HPLC using a variable wavelength UV detector operating at a  suitable
wavelength.  Chromatogram peaks are recorded and integrated using a re-
cording integrator. The concentration of the compound in the effluent from
the  generator column, i.e., the water solubility of the compound, is  deter-
mined from the mass of the compound (solute) extracted from a measured
volume of water (solvent).

     (ii)  Since the HPLC method is  only applicable to  compounds that
absorb in the UV, an alternate gas chromatographic (GC) method, or any
other reliable procedure (which must be approved by OPPTS), can be used
for  those compounds that  do  not absorb  in the UV. In the GC  method
the  saturated solutions produced in the  generator column  are extracted
using an appropriate organic solvent that is subsequently injected  into the
GC,  or  any  other suitable analytical device, for  analysis of the test
compound.

     (4)  Reference chemicals.  The  following table  1  lists  the  water
solubilities at 25 °C for a number of reference chemicals as obtained from
the  scientific literature. The data from Wasik et al. (see paragraph (e)(ll)
of this  guideline), Miller et al.  (see paragraph (e)(7) of this guideline),
and  Tewari et al. (see paragraph  (e)(10) of this guideline) were obtained
from the generator column method.  The water solubilities were also ob-
tained from Yalkowski et al. (see  paragraph (e)(13)  of this guideline),
Mackay et al. (see paragraph (e)(4) of this guideline) and other scientists
by the conventional shake-flask method. These data have been provided
primarily so that the generator column method can be calibrated from time
to time and to allow the chemical testing laboratory an opportunity to com-
pare its results with those listed in the following table 1. The water solu-

-------
      bility values at 25 °C reported by Yalkowski et al. (see paragraph (e)(13)
      of this guideline) are  their preferred values  and, in general, represent the
      best  available water  solubility data at  25 °C. The testing  laboratory  has
      the option of choosing its own reference chemicals, but references must
      be given to establish the validity of the  measured values of the water solu-
      bility and these chemicals must be approved by OPPTS.

              Table 1.—Water Solubilities at 25 tC of Some Reference Chemicals
Reference chemical
2-Heptanone
1-Chlorobutane 	
Ethylbenzene 	
1 ,2,3-Trimethylbenzene 	
Biphenyl
Phenanthrene
2 4 6-Trichlorobiphenyl
2,3,4,5-Tetrachlorobiphenyl 	
Hexachlorobenzene 	
2.3.4.5.6-PentachlorobiDhenvl 	
Water solubility (ppm at 25 °C)
Wasik (gen-
erator col-
umn meth-
od)
40802
8732
1872
65. 52
6.71310
1 .0024
0.2263 10
0.02093 10
0.005483 10
Yalkowski1
5
4300
872.9
208
75.2
7.48
1.212
0.225
0.01396
0.004669
0.004016
Other lit-
erature ref-
erences
43305
6667
1627
48.27
6.628
0.1198
0.01 928
0.009969
0.00688
  1 Preferred water solubility at 25 °C by Yalkowski et al. (1990) under paragraph (e)(13) of this guide-
line based on a critical review of all the experimental water solubility data published.
  2 Tewari et al. (1982) under paragraph (e)(10) of this guideline.
  3 Leifer et al. (1983) under paragraph  (e)(3) of this guideline.
  4 May, Wasik, and Freeman (1978, 1978a) under paragraphs (e)(5) and (6) of this guideline.
  5 Yalkowski et al. (1990) under paragraph (e)(13) of this guideline.
  6 Hansch et al. (1968) under paragraph (e)(2) of this guideline.
  7 Button and Calder (1975) under paragraph (e)(9) of this guideline.
  8 Mackay et al. (1980) under paragraph (e)(4) of this guideline.
  9 The elution chromatographic method from OECD (1981) under paragraph (e)(8) of this guideline.
  10 Miller et al. (1984) under paragraph (e)(7) of this guideline.

            (5) Applicability and  specificity, (i) Procedures are described in this
      test guideline to determine the water solubility for liquid or solid com-
      pounds. The water solubility can be determined in very pure water, buffer
      solution for compounds that reversibly ionize or protonate, or in artificial
      seawater as a function of  temperature  (i.e., in the range of temperatures
      of  environmental  concern). This guideline is not  applicable to the water
      solubility of gases.

            (ii) This  test guideline is  designed to determine the water solubility
      of  a solid  or liquid test chemical  in the range  of 1 ppb to  5000 ppm.
      For chemicals whose solubility is below 1 ppb,  the water solubility should
      be  characterized as "less than  1 ppb"  with no further quantification. For
      solubilities greater than 5000 ppm, the shake flask method under  OPPTS
      830.7840 should be used.
           (c) Test procedure—(1) Test  conditions—(i) Special  laboratory
       equipment. (A)  Generator column. Either of two  different designs shall

-------
be used depending on whether the eluted aqueous phase is analyzed by
HPLC under paragraph (c)(3)(ii) of this guideline or by solvent extraction
followed by GC (or any other reliable quantitative)  analysis of solvent
extract under paragraph (c)(3)(iv) of this guideline. The design of the gen-
erator column is shown in the following figure 1:

                   FIGURE 1—GENERATOR COLUMN
                                - GLASS WOOL
                                -SUPPORT (100-120 MESH
                                      CHROMOSORB W)
                                -6mm

                                -GLASS WOOL
The  column  consists of a 6 mm (V4-inch) O.D.  Pyrex tube joined to  a
short enlarged section of 9 mm Pyrex tubing which in turn is connected
to another section of 6 mm (V4-inch) O.D. Pyrex tubing.  Connections to
the inlet Teflon tubing (Vs-inch O.D.) and to the outlet stainless steel tub-
ing (Vie-inch O.D.) shall be made by means of stainless steel fittings with
Teflon ferrules. The column is enclosed in a water jacket for temperature
control as shown in the following figure 2:

-------
FIGURE  2—SETUP  SHOWING  GENERATOR COLUMN  ENCLOSED  IN A
WATER JACKET AND OVERALL ARRANGEMENT OF THE APPARATUS USED
IN THE GC METHOD
                                I-—GENERATOR
                                01   COLUMN
                                I I    INLET
                         BATH —
                         RETURN
                        COLLECTING-
                         VESSEL
                                    -GENERATOR
                                     COLUMN
                                     -TO CONSTANT
                                     TEMPERATURE
                                        BATH
                                   -EXTRACTING
                                     SOLVENT
                                    -WATER CONTAINING
                                       SOLUTE
     (B) Constant temperature bath with circulation pump-bath and capable
of controlling temperature  to  + 0.05 °C. (See  paragraph (c)(3) of this
guideline.)

     (C) High pressure  liquid chromatograph equipped with a variable
wavelenth UV absorption detector operating at a suitable wavelength and
a recording integrator under paragraph (c)(3)(ii) of this guideline.

     (D) Extractor column.  6.6 x 0.6 cm stainless steel tube with end fit-
tings containing  5 (im frits filled with a superficially porous phase packing
(Bondapack C7S/Corasil: Waters Associates) under paragraph (c)(3)(ii) of
this guideline.

     (E) Two 6-port high pressure  rotary switching valves under paragraph
(c)(3)(ii) of this guideline.

-------
     (F) Collection vessel.  An  8  x 3/4 in section of Pyrex tubing with a
flat bottom connected to a short  section of 3/s-in O.D. borosilicate glass
tubing in figure 2 under paragraph (c)(l)(i)(A) of this guideline. The col-
lecting vessel is sealed with a 3/s-in Teflon cap fitting under paragraph
(c)(3)(iii) of this guideline.

     (G) Gas chromatograph, or  any other reliable analytical equipment,
which has  a  detector sensitive to the solute of interest under paragraph
(c)(3)(iii) of this guideline.

     (ii) Purity of water. Water meeting ASTM Type II standards, or an
equivalent grade,  shall be used to minimize the  effects of dissolved salts
and other impurities on water solubility. ASTM Type II water is described
in ASTM  D 1193-77,  "Standard  Specification for  Reagent Water,"
ASTM D 1193-77. Copies of this  material may be  obtained from the
American Society  for Testing and Materials (ASTM), 1916  Race Street,
Philadelphia,  PA 19103.

     (iii) Purity of solvents. All solvents used in this  method shall  be
reagent or HPLC grade. Solvents shall contain no impurities  which could
interfere with the determination of the test compound.

     (iv) Seawater. When the water solubility in seawater is desired, the
artificial seawater described in  paragraph (c)(2)(ii) of this guideline shall
be used.

     (v) Effect of pH on solubility. For chemicals that reversibly ionize
or protonate  with a pKa  or pKb between 3  and  11,  experiments  shall  be
performed  at pH's 5.0,  7.0, and 9.0 (or any  other pH's  specified  by
OPPTS) using appropriate buffers.

     (2) Preparation of reagents and solutions—(i)  Buffer solutions.
Prepare buffer solutions as follows:

     (A) pH  3.0—To 250  mL  of 0.10M potassium hydrogen phosphate
add  111  mL  of 0.10 M hydrochloric  acid; adjust  the final volume  to
500 mL with  reagent grade water.

     (B) pH 5.0—To 250 mL of  0.1M potassium hydrogen phthalate add
113  mL of 0.1M sodium hydroxide; adjust the  final volume to 500 mL
with reagent grade water.

     (C) pH  7.0—To 250 mL  of 0.1M potassium dihydrogen phosphate
add  145 mL of 0.1M sodium  hydroxide;  adjust the  final volume  to
500 mL with  reagent grade water.

     (D) pH 9.0—To 250 mL of  0.075M borax add 69 mL of 0.1M HC1;
adjust the final volume to 500 mL with reagent grade water.

-------
           (E) pH  11.0—To 250 mL  of 0.05  M  sodium  bicarbonate  add
       113 mL of  0.10 M  sodium hydroxide;  adjust  the final  volume  to
       500 mL with reagent grade water.

       Check the pH of each buffer solution with a pH meter at 25 °C and adjust
       to pH 5.0, 7.0, or 9.0, if necessary. If the pH of the solution has changed
       by ±0.2 pH units or more  after the addition of the test compound, then
       a  more  concentrated buffer is required for that pH  determination.  The
       sponsor should then choose  a more suitable buffer.

           (ii) Artificial seawater. Add the  reagent-grade  chemicals  listed  in
       the following table 2 in the specified amounts and order to  890 mL  of
       reagent-grade water. Each chemical shall be dissolved before another one
       is added.

                      Table 2.—Constituents of Artificial Seawater1
Chemical
NaF 	
SrCI26H2O 	
H3BO3
KBr
KCI
CaCI22H2O
Na2SO4
MgCI26H2O
NaCI 	
Na2SiO39H2O 	
NaHCO, 	
Amount
3 mg
20 mg
30 mg
100 mg
700 mg
1 47 q
400 q
1078 g
23.50 g
20 mg
200 ma
  1 If the resulting solution is diluted to 1 cubic decimeter (1 L), the salinity should be 34+0.5 g/kg and
the pH 8.0+0.2. The desired test salinity is attained by dilution at time of use.

           (3) Performance of the test. Using either the procedures under para-
      graph (c)(3)(ii) or  (c)(3)(iii) of this guideline, determine the water solu-
      bility of the test compound at 25 °C in reagent grade water or buffer solu-
      tion,  as appropriate. Under certain circumstances, it may be necessary to
      determine  the  water solubility of a test compound at 25 °C in  artificial
      seawater. The water solubility can also be determined at other temperatures
      of environmental concern by adjusting the temperature of the water bath
      to the appropriate temperature.

           (i) Prior to the determination of the water solubility of the test chemi-
      cal, two procedures shall be followed.

           (A) The saturated aqueous solution leaving the generator column must
      be tested for the presence of an emulsion, using a Tyndall procedure. If
      colloids are present, they must be  eliminated prior to the injection into
      the extractor  column. This  may  be achieved by lowering the flow rate
      of the water.

                                         8

-------
     (B)  The efficiency of the removal of the solute (i.e. test chemical)
by the solvent extraction from the extraction column must be determined
and used in the determination of the water solubility of the test chemical.

     (ii)  Procedure A—HPLC Method—(A)  Scope.  (7)  Procedure A
covers the determination of the aqueous  solubility of compounds which
absorb in the UV.  The HPLC analytical system is  shown  schematically
in the following figure 3:

 FIGURE 3—SCHEMATIC OF HPLC—GENERATOR COLUMN FLOW SYSTEM
                        GENERATOR COLUMN
             SAMPLE INJECTION VALVE
                   SAMPLE LOOP
                               EXTRACTOR COLUMN
                                                             WASTE
Two reciprocating piston pumps  deliver the mobile phase (water or  sol-
vent/water mixture) through  two 6-port high  pressure rotary valves  and
a 30 x 0.6 cm C7S/Corasil analytical column to a variable wavelength UV
absorption detector operating  at a suitable wavelength;  chromatogram
peaks are recorded and integrated with  a recording integrator. One of the
6-port valves is the sample injection valve used for injecting samples of
standard solutions of the solute in an appropriate concentration for deter-
mining response factors of standard solutions of basic chromate for deter-
mining  the  sample  loop volume. The  other 6-port valve in the system
serves as  a switching  valve for the  extractor column which is used to re-
move solute from the aqueous solutions.

     (2) The general procedure for analyzing the aqueous phase is as  fol-
lows (a detailed  procedure is given in paragraph (c)(3)(ii)(B)(/) of  this
guideline).

     (/)  Direct the aqueous solution to "Waste" (see figure 3 under para-
graph (c)(3)(ii)(A)(7)  of this guideline) with  the  switching  valve in the
inject position in order to  equilibrate internal surfaces with  the solution,
thus ensuring that the analyzed sample would not be  depleted by solute
adsorption on surfaces upstream from the valve.

-------
     (//) At the  same time, water is  pumped from the HPLC pumps in
order to displace the solvent from the extractor column.

     (///) The switching valve is next changed to the load position to divert
a sample of the solution through the extractor column, and the liquid leav-
ing this column is collected  in a weighing bottle. During  this extraction
step, the mobile phase is changed to a solvent/water mixture to  condition
the analytical column.

     (iv) After the desired volume  of sample is  extracted, the  switching
valve is returned to the inject position for elution and analysis. Assuming
that there is no breakthrough of solute from the  extractor  column during
the extraction step, the chromatographic  peak represents all of the solute
in the  sample, provided  that the extraction efficiency is  100 percent. If
the extraction efficiency is less than 100 percent, then the extraction effi-
ciency shall be used to determine the actual weight of the solute extracted.

     (v) The solute concentration in the aqueous phase is calculated from
the peak area and the weight of the extracted liquid collected in the weigh-
ing bottle.

     (B) Determinations—(7) Sample Loop Volume. Accurate measure-
ment  of  the   sample  loop  may  be   accomplished  by  using  the
spectrophotometric method of Devoe et  al. (see  paragraph (e)(l) of this
guideline). For this method measure  absorbance, Aioop, at 373  nm of at
least three solutions,  each of which  is prepared by  collecting  from the
sample  valve an appropriate  number, n,  of loopfuls of an aqueous stock
solution of K.2CrO4 (1.3  percent by weight) and diluting to 50 mL with
0.2 percent KOH. (For a 20 (iL loop, use n = 5; for a 50 |iL  loop, use
n = 2.) Also measure the absorbance, Ast0ck, of the  same stock solution
after diluting 1:500 with 0.2  percent KOH. Calculate the loop volume to
the nearest 0.1 [iL using the equation:

                    Vioop =(Aioop/Astock)(10-4/n)

     (2) Response Factor (RF). (/) For all determinations  adjust the mo-
bile phase solvent/water ratio and flow rate  to obtain a reasonable retention
time on the HPLC column. For example, typical concentrations of solvent
in the mobile phase range from 50  to  100  percent while flow rates range
from  1  to  3 mL/min; these conditions give a 3 to 5  min retention time.

     (//) Prepare standard solutions of known concentrations of the solute
in a suitable solvent. Concentrations must give a recorder response within
the maximum response of the  detector.  Inject samples of each standard
solution into the HPLC system using the calibrated sample loop. Obtain
an average peak area from at  least three injections of each standard sample
at a set absorbance unit full scale  (AUFS), i.e.,  at the same absorbance
scale attenuation setting.

                                 10

-------
     (///) Calculate the response factor from the following equation:
                                       Concentration (M)
            Response Factor (RF) =
                                    (Average Area) (AUFS)

        Loading of the Generator Column. (/) The design of the genera-
tor  column was described in paragraph (c)(l)(i)  of this guideline and is
shown in figure 1 under paragraph (c)(l)(i)(A) of this guideline. To pack
the  column, a  plug of silanized glass wool is inserted into one  end of
the  6 mm Pyrex tubing. Silanized diatomaceous silica support (about 0.5g
100-120 mesh Chromosorb  W  chromatographic   support  material) is
poured  into the tube with tapping and retained with  a second plug of
silanized glass wool.

     (//) If the solute is a liquid, the column is loaded by pulling the liquid
solute through the dry support with gentle suction. If the solute is a solid,
a 1  percent solution of the solid in a volatile solvent is added to the dry
packing. The solvent  is then distilled off the column under reduced pres-
sure. After loading the  column draw  water up through  the column to re-
move entrapped air.

     (4) Analysis of  the solute. Use the following procedure to collect
and analyze the solute.

     (/) Pump water  to the  generator column by means of  a minipump
or pressurized water reservoir as  shown in the following figure 4:
                                 11

-------
            FIGURE 4—WATER RESERVOIR FOR GC METHOD
                          'TO
                        COMPRESSED
                        GAS CYLINDER
                                      TO GENERATOR
                                      COLUMN INLET
With the switching valve (figure 3 under paragraph (c)(3)(ii)(A)(7) of this
guideline) in the inject position (i.e., water to waste), pump water through
the generator column at a flow rate of approximately 1 mL/min for ap-
proximately 5 minutes to bring the system into equilibrium.

    (//) Flush out the solvent  that remains in the system from previous
runs by changing the mobile phase to  100 percent H2O and allowing the
water to reach the HPLC detector, as indicated by a negative reading. As
soon as this occurs, place a 25 mL weighing bottle (weighed to the nearest
milligram) at the waste position and immediately turn the  switching valve
to the load position.

    (///) Collect an  amount of water  (as determined by  trial and  error)
in the  weighing bottle, corresponding  to the amount of solute  adsorbed
by the  extractor column that gives a large on-scale detector response. Dur-
ing this extraction step,  switch back to the original HPLC mobile  phase
composition, i.e.,  solvent/water mixture, to condition the HPLC analytical
column.

    (iv) After the desired volume of sample has been extracted, turn the
switching  valve back  to the inject position (figure 3 under  paragraph
                                 12

-------
(c)(3)(ii)(A)(7) of this guideline); at the  same time turn on the recording
integrator. The solvent/water mobile phase will elute the solute from the
extractor column and transfer the solute to the HPLC analytical column.

     (v) Remove the weighing bottle, cap it and replace it with the waste
container. Determine the weight of water collected to the nearest milligram
and  record the corresponding peak  area. Using the  same AUFS setting
repeat the analysis of the solute at least 2 more times  and determine the
average ratio of peak area to grams of water collected. Calculate the solute
solubility in water using the following equation:

                     s = (997 g/L)(RF)(Vioop)(AUFS)(R)

where
     s = solubility (M)

     RF = response factor

     Vioop = sample loop volume (L)

     R = ratio of area to grams of water.

     (iii) Procedure B—GC Method. (A) Scope. In the GC method, or
any other analytical method, aqueous solutions from the generator column
enter a  collecting vessel (figure 2  under paragraph (c)(l)(i)(A) of this
guideline) containing  a known weight of extracting solvent which  is im-
miscible in water. The  outlet  of the generator column  is positioned such
that  the aqueous  phase always enters below the extracting solvent. After
the aqueous phase is  collected, the  collecting vessel is  stoppered and the
quantity of  aqueous phase is  determined by weighing. The solvent and
the aqueous  phase are equilibrated by slowly rotating the collecting vessel.
The  extraction efficiency of the  solvent  must be determined at this time.
A small amount  of the extracting solvent is  removed and injected into
a gas chromatograph equipped with an appropriate detector. The  solute
concentration in the aqueous phase is determined from a calibration curve
constructed using known concentrations of the solute.

     (B) Alternative Method. If another (approved) analytical method is
used instead of the  GC,  that method shall be  used to determine quan-
titatively the amount of solute present in the extraction solvent.

     (C) Determinations—(7) Calibration curve. (/) Prepare solute stand-
ard solutions of concentrations covering the range of the solute solubility.
Select a column and optimum GC operating conditions for resolution be-
tween the solute and  solvent and the solute and extracting solvent. Inject
a known volume of each standard solution into the injection port  of the
GC.  For each standard solution determine the average of the ratio R of
peak area to volume (in microliters) for three chromatographic peaks from
three injections.

                                 13

-------
     (//)  After  running all the  standard  solutions,  determine the coeffi-
cients, a and  b,  using  a  linear  regression  equation of  concentration
(C) vs. R in the following form

                                C = aR + b

     (///) If another analytical method  is used, the procedures described
in paragraph (c)(3)(iii)(C)(7) of this guideline shall be used to determine
quantitatively the amount of solute in the extraction solvent.

     (2)  Loading of the generator column.  The generator column  is
packed and loaded with solute in the  same manner as for the HPLC meth-
od  (see paragraph (c)(3)(ii)(C)(3) of this guideline). As shown in figure
2 under  paragraph (c)(l)(i)(A) of this guideline, attach approximately 20
cm of straight stainless steel tubing to the bottom of the generator column.
Connect  the top  of the  generator column to a water  reservoir  (figure 4
under paragraph (c)(3)(ii)(C)(/)(/) of this guideline) using Teflon tubing.
Use air or nitrogen  pressure (5 PSI) from an air or nitrogen cylinder to
force water from the reservoir  through the  column. Collect water in an
Erlenmeyer flask for approximately 15 min while the solute  concentration
in water equilibrates; longer time may be required for less  soluble com-
pounds.
        Collection and extraction of the solute. During the equilibration
time, add a known weight of extracting solvent to a collection vessel which
can be capped. The extracting solvent should cover the bottom of the col-
lection vessel to a depth sufficient to submerge the collecting tube but
still maintain  100:1 water/solvent ratio. Record the weight (to the  nearest
milligram) of a collection vessel with cap and extracting solvent. Place
the collection vessel under the  generator column  so that water from the
collecting tube enters below the level of the extracting solvent (figure 2
under paragraph (c)(l)(i)(A) of this guideline). When the collection vessel
is  filled, remove it from under the  generator  column, replace cap, and
weigh the filled vessel. Determine the weight of water collected.  Before
analyzing for the  solute,  gently shake the  collection vessel  contents for
approximately 30 min,  controlling the rate of shaking so as not to  form
an emulsion; rotating the flask end over end 5  times per minute is suffi-
cient.

    (4) Analysis of the solute.  (/) After shaking, allow the collection ves-
sel to stand for approximately 30 min; then remove a known volume of
the extracting solvent from the vessel using a microliter syringe and inject
it into the GC. Record the ratio  of peak area to volume injected and,  from
the regression equation of the calibration line, determine the concentration
of solute in the extracting  solvent. The concentration  of solute in water
C(M) is determined from the following equation:

                  C(M) = (Ces) [(dH20/des)][(ges/gH2o)]

                                  14

-------
where

     Ces is the concentration of solute in extracting solvent (M),

     dmo and des are the densities of water and extracting solvent, respec-
tively, and

     ges and gH2o are the grams of extracting solvent and water, respec-
tively, contained in the collection vessel.

     (//) Make replicate injections from each collecting vessel to determine
the average solute concentration in water for each vessel. To make sure
the generator column has reached equilibrium, run at least two additional
(for a total  of three) collection vessels and analyze the  extracted solute
as described above.  Calculate the water solubility of the solute from the
average solute concentration in the three vessels.

     (iv) Modification of procedures for potential  problems. If the test
compound decomposes in one or more of the aqueous solvents required
during the period of the test at a rate such that an accurate value for water
solubility cannot be obtained, then it will be necessary to carry out detailed
transformation  studies; e.g.,  hydrolysis under OPPTS 830.2110. If decom-
position is due to aqueous photolysis, then it will be necessary to carry
out water solubility  studies in the dark, under red or yellow lights, or by
any other suitable method to eliminate this transformation process.

     (d) Data  and reporting—(1) Test report,  (i) For each set of condi-
tions, (e.g.,  temperature,  pure water, buffer solution, artificial  seawater)
required for the study, provide the water solubility value for each of three
determinations, the mean value, and the standard deviation.

     (ii) For compounds that decompose at a rate such that a precise value
for the water  solubility cannot  be obtained,  provide  a statement to that
effect.

     (iii) For compounds with water solubility below 1  ppb, report  the
value as "less than 1 ppb".

     (2) Specific analytical, calibration and recovery procedures, (i) For
the HPLC method describe and/or report:

     (A)  The method used to  determine the sample  loop  volume  and the
average and standard deviation of that volume.

     (B) The average and standard deviation  of the response factor.

     (C) Any changes made  or problems encountered in the test procedure.

     (ii) For the GC, or any other analytical, method report:

                                  15

-------
     (A) The column and GC operating conditions of temperature and flow
rate, or the operating conditions of any other analytical method used.

     (B) The average and  standard deviation of the  average  area per
microliter obtained for each of the standard solutions.

     (C) The form of the regression equation obtained in the calibration
procedure.

     (D) The extracting solvent used, and its extraction efficiency.

     (E) The average and standard deviation of solute concentration in
each collection vessel.

     (F) Any changes made or problems encountered in the test procedure.

     (G) If applicable, a  complete description of the analytical method
which was used instead of the GC method.

     (e) References. The following references should be consulted for ad-
ditional background material on this test guideline.

     (1) DeVoe, H.  et al.,  Generator  columns and high pressure liquid
chromatography for determining aqueous  solubilities and octanol-water
partition coefficients of hydrophobic substances. Journal of Research, Na-
tional Bureau of Standards, 86:361-366 (1981).

     (2) Hansch, C. et al.,  The linear free-energy relationship between par-
tition coefficients,  and the aqueous solubility of organic liquids. Journal
ofOranic  Chemistry 33:347-350 (1968).

     (3) Leifer, A. et al.,  Environmental transport and transformation of
polychlorinated  biphenyls.  Chapter  1.  U.S. Environmental Protection
Agency Report: EPA-560/5-83-005 (1983).

     (4) Mackay, D. et al.,  Relationship between aqueous solubility and
octanol-water partition coefficient. Chemosphere 9:701-711 (1980).

     (5) May, W.E. et al., Determination of the aqueous  solubility  of
polynuclear  aromatic  hydrocarbons   by  a  coupled  column  liquid
chromatographic technique. Analytical Chemistry 50:175-179 (1978).

     (6) May, W.E. et al. Determination of the solubility behavior of some
polycyclic aromatic hydrocarbons  in  the water.  Analytical  Chemistry,
50:997-1000 (1978a).

     (7) Miller, N.M. et al., Aqueous  solubilities,  octanol/water partition
coefficients, and entropy of melting of chlorinated benzenes and biphenyls.
Journal of Chemical and Engineering Data 29:184-190 (1984).

                                  16

-------
    (8) OECD/Organization for Economic Cooperation and Development.
Test Guideline No. 105. Water solubility column elution-flask method
(1981).

    (9) Sutton, C. and Calder, J.A., Solubility of alkylbenzenes in distilled
water and seawater at 25 °C. Journal of Chemical and Engineering Data
20:320-322 (1975).

    (10) Tewari , Y.B. et al., Aqueous solubility  and octanol/water parti-
tion coefficient of organic compounds at 25 °C. Journal of Chemical and
Engineering Data 27:451-454 (1982).

    (11) Wasik, S.P. et al., Octanol/Water Partition Coefficient and Aque-
ous Solubilities of Organic Compounds. NBS Report NBSIR  81-2406.
Washington, DC: National Bureau of Standards, U.S. Department of Com-
merce (1981).

    (12) Wasik. S.P. et al., Water solubility  and octanol/water partition
coefficient  of polychlorinated  biphenyls and  other selected substances.
Task 1C. Interagency Agreement EPA-80-D-X0958 between the U.S. En-
vironmental Protection Agency and the National Bureau of Standards
(1982).

    (13) Yalkowski, S.H. et al., "Arizona database of aqueous solubilities
of organic compounds"; Fifth Edition. University of Arizona, College of
Pharmacy, Tucson, AZ 85721 (1990).
                                17

-------