www.epa.gov/or
United States
Environmental Protection
Agency
             Surface Analysis of Nerve Agent
             Degradation Products by Liquid
             Chromatography/Tandem Mass
             Spectrometry (LC/MS/MS)

Office of Research and Development
National Homeland Security Research Center

-------
                                                                       EPA/600/R-13/224
                                                                       September 2013
SURFACE ANALYSIS OF NERVE AGENT DEGRADATION PRODUCTS BY LIQUID
      CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY (LC/MS/MS)

                 Sampling and Analytical Method for Wipe Analysis of Surfaces
                                       Revision 1

                       United States Environmental Protection Agency
                         National Homeland Security Research Center
                             26 W. Martin Luther King Jr. Drive
                                  Cincinnati, OH 45268

                                          and

                         Centers for Disease Control and Prevention
                     National Institute for Occupational Safety and Health
                                     5555 Ridge Ave
                                  Cincinnati, OH 45213

                                   Last Revised: 12/12

-------
DISCLAIMER

The United States Environmental Protection Agency through its Office of Research and Development
(ORD), National Homeland Security Research Center (NHSRC), funded and managed the research
described here (IA #DW-75-922440001-0) in collaboration with the National Institute of Occupational
Safety and Health (NIOSH), Centers for Disease Control and Prevention (CDC), a division of the U.S.
Department of Health and Human Services (DHHS). It has been subjected to the Agency's administrative
review and approved for publication. The views expressed in this paper do not necessarily reflect the
views or policies of the Agency.  Mention of trade names or commercial products does not constitute
endorsement or recommendation for use.
Questions concerning this document or its application should be addressed to:

Stuart Willison, Ph.D.
Project Officer
U.S. Environmental Protection Agency
National Homeland Security Research Center
26 W. Martin Luther King Drive, MS NG16 Cincinnati, OH 45268
513-569-7253
Willison.Stuart@epa.gov

Robert Streicher, Ph.D.
Project Officer
National Institute for Occupational Safety and Health Laboratories
Alice Hamilton Laboratory
5555 Ridge Avenue
Cincinnati, OH 45213
513-841-4296
Rps3@cdc.gov

-------
ACKNOWLEDGMENTS

We would like to acknowledge the following individuals and organization for their contributions towards
the development and/or review of this method.

United States Environmental Protection Agency (EPA)
       Office of Research and Development, National Homeland Security Research Center
       Stuart Willison, Project Officer and Method Development
       Matthew Magnuson, Technical Reviewer

United States Environmental Protection Agency (EPA)
       Office of Emergency Management
       Terry Smith, Technical Reviewer

United States Environmental Protection Agency (EPA)
       Office of Ground Water and Drinking Water
       April Dupre, Technical Reviewer

Centers for Disease Control and Prevention
       National Institute for Occupational Safety and Health
       Jack Pretty, Laboratory Advisor
       Robert Streicher, Project Officer

-------
                                   EXECUTIVE SUMMARY
The  sampling and analytical method described herein  was  developed and  tested within the same
laboratory to assess the recoveries of nerve agent degradation products from various porous (vinyl tile,
painted drywall, wood) and mostly nonporous (laminate, galvanized steel, glass) surfaces. Performance
data (method detection limit and precision and accuracy data) are available to demonstrate the fitness-for-
purpose regarding the development  of a method for nerve agent degradation products in that single
laboratory.  Samples  are collected from  surfaces using wipes, the wipes are  spiked with a  surrogate
compound and carried through extraction with distilled water by sonication and filtration steps followed
by analysis  using liquid chromatography electrospray ionization/tandem mass spectrometry  (LC/ESI-
MS/MS) by  direct injection without derivatization. Detection limit data were generated using wipes on a
laminate surface following the procedures of 40 CFR Part 136, Appendix B, as  part of EPA's guidelines
for determining a method detection limit.

Gauze wipes were selected  over other tested  wipes (i.e.,  filter paper, glass fiber filters, nonwoven
polyester fiber) because gauze wipes  were physically robust during the wiping procedure, contained low
background  levels, produced no peaks that interfered with the target analytes, and produced  the highest
percent recoveries of all wipes tested during sample analysis.  Percent recoveries were highest for the
laminate surface and ranged from 65-87 % for all of the nerve agent degradation  products analyzed in ESI
negative mode.  The  resulting equivalent method detection limits obtained from wiping the laminate
surface were  0.04  |o,g/cm2 for  isopropyl   methylphosphonic   acid  (IMPA),  0.07  |o,g/cm2  for
methylphosphonic acid (MPA), 0.05  |o,g/cm2 for ethyl methylphosphonic acid (EMPA), 0.07 |o,g/cm2 for
ethyl hydrogen  dimethylamidophosphate,  sodium  salt (EHDMAP)  and 0.02  |o,g/cm2 for  pinacolyl
methylphosphonic acid (PMPA).  Diisopropyl methylphosphonate (DIMP) was not recovered  unless the
surfaces were wiped  immediately after spiking due to the volatile nature of this compound.  Other
complications are presented in the method in section 14.4.  Precision and accuracy data were generated
from each tested surface fortified with these analytes.
                                               IV

-------
     SURFACE ANALYSIS OF NERVE AGENT DEGRADATION PRODUCTS BY LIQUID
          CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY (LC-MS/MS)

                            TABLE OF CONTENTS

      SECTION                                                   PGNO.
      DISCLAIMER                                                ii

      ACKNOWLEDGMENTS                                          iii

      EXECUTIVE SUMMARY                                        iv

      LIST OF TABLES                                              vii

      LIST OF ACRONYMS AND ABBREVIATIONS                          viii

1.     INTRODUCTION                                              1

2.     SCOPE AND APPLICATION                                      1

3.     SUMMARY OF METHOD                                        3

4.     DEFINITIONS                                                4

5.     INTERFERENCES                                              5

6.     HEALTH AND SAFETY                                          6

7.     EQUIPMENT AND SUPPLIES                                     6

8.     REAGENTS AND STANDARDS                                    7

9.     SAMPLE COLLECTION, PRESERVATION AND STORAGE                 9

10.    QUALITY CONTROL                                           10

11.    INSTRUMENT CALIBRATION AND STANDARDIZATION                 15

12.    ANALYTICAL PROCEDURE                                      16

13.    DATA ANALYSIS AND CALCULATIONS                             18

14.    METHOD PERFORMANCE                                       18

15.    POLLUTION PREVENTION                                       19

16.    WASTE MANAGEMENT                                        19

17.    REFERENCES                                                21

18.    TABLES AND VALIDATION DATA                                 22

19.    ATTACHMENTS                                              28

-------
LIST OF TABLES

Table 1. Materials Tested for the Wipe Analysis of Nerve Agent Degradation Products	2
Table 2. Holding Time Sample Stability of Nerve Agent Degradation Analytes of Wipe Samples in ESI Negative
Mode	2
Table 3. Method Parameters for Nerve Agent Degradation Products	3
Table 4a. ESI (+) MRM Ion Transitions, Retention Time (RT) and Variable Mass Spectrometer Parameters	4
Table 4b.  ESI (-) MRM Ion Transitions, Retention Time (RT) and Variable Mass Spectrometer
Parameters	 25
Table 5. ESI (+) and (-) MS/MS Conditions	5
Table 6. Liquid Chromatography Gradient Conditions	5
Table 7. Target Concentrations of Calibration Standards Used During the Development of this Method (ng/mL)	6
                                                 VI

-------
LIST OF ACRONYMS AND ABBREVIATIONS
   AS        Analyte Stock Standard (Solution)
   CAL      Calibration Standard
   CAS      Chemical Abstracts Service
   CCC      Continuing Calibration Check
   CDC      Centers for Disease Control and Prevention
   CID       Collisionally Induced Dissociation
   CWA      Chemical Warfare Agent
   DL        Detection Limit
   DHHS     U.S. Department of Health and Human Services
   DIMP     Diisopropyl Methylphosphonate
   DQO      Data Quality Objective
   EHDMAP  Ethyl Hydrogen Dimethylamidophosphate, sodium salt
   EMPA     Ethyl Methylphosphonic Acid
   EPA      U.S. Environmental Protection Agency
   ERLN     Environmental Response Laboratory Network
   ESI (+)     Electrospray lonization in Positive Mode
   ESI  (-)     Electrospray lonization in Negative Mode
   FD        Field Duplicate
   IDC       Initial Demonstration of Capability
   IDL       Instrument Detection Limit
   IMPA     Isopropyl Methylphosphonic Acid
   LC        Liquid Chromatography
   LC/MS/MS Liquid Chromatography Coupled with Tandem Mass Spectrometry
   LFB      Laboratory Fortified Blank
   LFSM     Laboratory Fortified Sample Matrix
   LFSMD   Laboratory Fortified Sample Matrix Duplicate
   LMB      Laboratory Method Blank
   MDL      Method Detection Limit
   MPA      Methylphosphonic Acid
   MRL      Minimum Reporting Limit
   MRM     Multiple Reaction Monitoring
   MS        Mass Spectrometer(try)
   MSDS     Material Safety Data Sheet
   MS/MS    Tandem Mass Spectrometry
   NHSRC   National Homeland Security Research Center
   NIOSH    National Institute for Occupational Safety and Health
   NIST      National Institute of Standards and Technology
   ORD      U.S. EPA's Office of Research and Development
   OSHA     Occupational Safety and Health Administration
   PIR       Prediction Interval of Result
   PMPA     Pinacolyl Methylphosphonic Acid
   ppb        Parts Per Billion
   ppm      Parts Per Million
   P&A      Precision and Accuracy
   PVDF     Polyvinylidene Fluoride
   QC        Quality Control
   r2         Coefficient of determination
                                             VII

-------
REC       Percent Recovery
RL        Reporting Limit
RPD       Relative Percent Difference
RSD       Relative Standard Deviation
RT        Retention Time
SAM      Selected  Analytical Methods for  Environmental Restoration  Following  Homeland
           Security Events
SD        Standard Deviation
S/N        Signal to Noise
SS         Surrogate Standard
SSS        Stock Standard Solution
VOA      Volatile Organic Analysis
X        Average Percent Recovery
a          Standard Deviation
                                          VIM

-------
1.  INTRODUCTION

    1.1. The  U.S. Environmental Protection Agency (EPA) is  responsible for developing tools  and
        methodologies which will enable the rapid  characterization of indoor and outdoor areas  and
        water systems following a deliberate/accidental release or a natural disaster.  EPA's National
        Homeland  Security Research  Center (NHRSC),  published Selected Analytical Methods for
        Environmental Remediation and Recovery (SAM),  formerly referred to as the Standardized
        Analytical Methods for Environmental Restoration  Following Homeland Security  Events  (1),
        which  is a compendium of methods that informs sample  collection  and analysis during the
        response to an all-hazards incident.   Chemical warfare  agents (CWAs) and their degradation
        products remain  a high-priority concern due to the potential for the intentional or unintentional
        release of these  agents. Nerve agents are very dangerous CWAs, which can break down into
        degradation products sufficiently persistent and toxic to be of interest during site remediation
        after a release.  Accordingly,  if an incident were to occur, versatile sampling procedures are
        needed to detect CWA degradation products from various CWAs and help determine the spread
        and concentration of these agents and degradation  products in contaminated areas.  Multiple
        types of contaminated  surfaces from an indoor setting (e.g., walls, posts, windows, floors  and
        furniture) will need to be extensively tested within  the contaminated  areas.  Direct extraction
        may be a possibility; however, the laboratory procedures can be tedious, complex,  and require
        the destruction of the material being analyzed.  Wipe sampling is preferred  because it can be
        performed  quickly and easily in a manner less destructive to the tested surface when  direct
        extraction is not feasible.

    1.2. After sample collection, selective analysis methods must be implemented to detect and quantify
        the appropriate agent and/or degradation products in the environmental sample. The appropriate
        procedure should account for possible contaminants already present within the sample as well as
        other matrix complications that may arise during analysis to ensure sample integrity and to
        ensure that the analysis method is applicable to the  matrix of interest. Liquid chromatography-
        tandem mass spectrometry (LC-MS/MS) is often the most appropriate and powerful analysis
        technique for polar nonvolatile compounds. LC-MS/MS affords laboratories an enhanced
        capability to analyze specific environmental matrices for CWA degradation products while
        avoiding complications that may arise from derivatization, a step more commonly  needed for gas
        chromatography/mass spectrometry analysis.  Although LC-MS analysis methods do exist for
        nerve agent degradation products from water, currently no known wipe sample collection and
        analysis protocol for the detection of nerve agent degradation products from contaminated
        surfaces is documented in the scientific literature.

2.  SCOPE AND APPLICATION

       2.1.    This sampling and analytical procedure was developed and tested in the same laboratory
               to investigate nerve agent degradation products, which may persist at a contaminated site,
               via  surface  wiping  followed  by  analytical  characterization.    The performance data
               presented demonstrate  the fitness-for-purpose regarding surface analysis  in that  single
               laboratory.   Surfaces (laminate, glass, galvanized steel, vinyl  tile, painted drywall  and
               treated  wood) were wiped with cotton  gauze wipes, sonicated, extracted with distilled
               water, and filtered. Samples were analyzed with direct injection electrospray ionization
               liquid  chromatography   tandem  mass  spectrometry   (ESI-LC-MS/MS)   without
               derivatization.   Detection limit data  were generated for all analytes of interest on a
               laminate surface. Accuracy and precision data were generated from each surface fortified
               with these analytes. The following analytes have been determined using this procedure:

-------
    Analyte                                                CAS Registry Number
       Diisopropylmethylphosphonate (DIMP)                              1445-75-6
       Ethyl Hydrogen Dimethylamidophosphate, sodium salt (EHDMAP)     2632-86-2
       Ethyl Methylphosphonic acid (EMPA)                               1832-53-7
       Isopropyl Methylphosphonic acid (IMPA)                            1832-54-8
       Methylphosphonic acid (MPA)                                      993-13-5
       Pinacolyl Methylphosphonic Acid (PMPA)                           616-52-4

2.2.    Wipe sampling can be performed quickly and easily when direct extraction is not feasible
       (e.g., walls, posts, windows, floors and furniture) as wipe sampling can be performed
       without the destruction of the tested surface.  Porous surfaces may have lower recoveries
       and less precision because the contaminants may sorb into the material.  Wipe sampling
       will  recover analyte only from the surface of the analyzed material.  It is, therefore,
       important to understand wipe efficiencies and the materials being wiped. This procedure
       assesses the recoveries from several porous and nonporous surfaces using wipes.

2.3.    Method detection limit (MDL) metrics are presented using EPA conventions (2-3).  The
       detection limit is defined as the statistically calculated minimum concentration that can
       be measured with 99% confidence  that the reported value is greater than zero (4).  The
       MDL  is  compound-dependent  and  reliant  on sample  preparation, sample  matrix,
       concentration  and instrument performance.   The  statistical procedure,  utilizing the
       Laboratory Fortified Sample Matrix samples (LFSM) and LFSM duplicates (LFSMDs),
       is used to calculate recovery.  Precision and accuracy (P&A) studies are performed as an
       initial  demonstration  of capability (IDC) and ongoing demonstration of capability to
       perform the procedure, including changes in  instrumentation  and operating conditions.
       These  studies evaluate whether the  reporting limits (RLs)  and calibration  standard
       concentrations are appropriate.

2.4.    This procedure is intended for use  by analysts skilled in the operation of LC-MS/MS
       instrumentation  and the  interpretation of the associated  data.   Due to the inherent
       complexities of LC-MS/MS  analysis, including the need to relate sample characteristics
       to  analytical  performance,  laboratories  should  update  their initial   estimates of
       performance and should strive to tighten their quality control limits as more experience is
       gained with this particular procedure.

2.5.    METHOD FLEXIBILITY

       Many variants of liquid chromatography (LC)  and Tandem Mass Spectrometry (MS/MS)
       technology are currently  in  operation. In addition, variability exists in the sources of
       wipe materials, wipe composition, and compatibility of various wipe materials with some
       surfaces.  This procedure was developed using  a triple quadrupole LC-MS/MS,  with
       optimized LC conditions and wipe materials.  The procedure has been verified using only
       the specified equipment and  conditions. Other types of LC-MS/MS instrumentation, LC
       and/or  ESI-MS/MS  conditions,   sample  collection   and  processing  steps,  and
       wipe/collection materials can be used for analysis as long  as similar performance  is
       demonstrated and the quality control measures outlined in section 10 of this report are
       implemented.

-------
3.  SUMMARY OF METHOD
       3.1.    Samples are collected from surfaces with wipes and stored at 4 °C (± 2 °C) if samples are
               not to  be analyzed within  a 24-hour time period.   When the samples are analyzed,
               samples are spiked with the appropriate surrogate compounds,  the appropriate solvent
               volume is added, the sample solution is sonicated, extracted with a syringe filter unit,
               then the extract is analyzed directly by LC-MS/MS operated simultaneously in positive
               and negative  electrospray  ionization modes, (ESI+) and  (ESI-)  respectively.   Data
               described in this procedure refer to ESI (-) mode because some complications can occur
               in ESI (+) mode.

       3.2.    Each target compound is separated chromatographically and identified by retention time.
               Comparison of the sample primary multiple reaction monitoring (MRM) transition to the
               known standard MRM transition from reference spectra under identical LC-MS/MS
               conditions is used to identify analytes. The retention time for the analytes of interest must
               fall within the retention time window of the standard  (within ±5%). The concentration
               of each analyte  is determined by the instrumentation software using external calibration.
               Surrogate analytes are added to samples to monitor extraction efficiency of the method
               analytes from the wipe and extraction process.

       3.3.    This procedure utilizes cotton gauze wipes, which were determined to provide the highest
               recoveries with  the least interference for any targeted analyte. Other wipes such as filter
               paper or glass fiber filters did have comparable recoveries and might be an appropriate
               alternative but  would not  be as robust during the wiping procedure for  the  targeted
               analytes.

-------
4.   DEFINITIONS

   4.1.  ANALYSIS BATCH - A set of samples analyzed on the same instrument within a 24-hour
        period and including no more than 20 field samples, beginning and ending with the analysis of
        the appropriate continuing calibration check (CCC) standards. Additional CCCs may be required
        depending on the number of samples (excluding QC samples) in the analysis batch and/or the
        number of field samples.

   4.2.  CALIBRATION STANDARD (CAL) - A solution prepared from the analyte stock standard
        solution and the surrogate/internal standard(s). The CAL solutions are used to calibrate the
        instrument response with respect to analyte concentration.

   4.3.  COLLISIONALLY INDUCED DISSOCIATION (CID) - The process of converting the
        precursor ion's translational energy into internal energy by collisions with neutral gas molecules
        to bring about dissociation into product ions.

   4.4.  CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing the method
        analytes and surrogate standard(s). The CCC is analyzed periodically to verify the accuracy of
        the existing calibration for those analytes at or near the mid-level concentrations. Low
        calibration concentrations can be added, in addition to mid-level concentrations, for further
        accuracy, but are not required.

   4.5.  DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be identified,
        measured, and reported with 99% confidence that the analyte concentration is greater than zero.

   4.6.  EXTRACTION BATCH - A set of up to twenty field samples (excluding quality control [QC]
        samples) extracted together using the same solvents, surrogate(s), fortifying solutions, and
        sampling devices.

   4.7.  FIELD DUPLICATE (FD) - Separate samples collected at the same time and place, under
        identical circumstances and treated exactly the same as other field samples throughout field
        and/or laboratory procedures. Analyses of FDs will give a measure of the precision associated
        with sample collection, preservation, and storage, as well as laboratory procedures.

   4.8.  LABORATORY FORTIFIED BLANK (LFB) - A blank matrix to which known quantities of
        the method analytes are added in the laboratory. The LFB is analyzed exactly like a sample, and
        its purpose is to demonstrate that the methodology is in control and that the laboratory is capable
        of making accurate and precise measurements.

   4.9.  LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - A field sample to which known
        quantities of the method analytes are added in the laboratory. The LFSM is processed and
        analyzed exactly like a sample, and its purpose is to determine whether the sample matrix
        contributes bias to the  analytical results. The background concentrations of the analytes in the
        sample matrix must be determined in a separate  sample.

   4.10. LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A duplicate of the
        field sample used to prepare the LFSM. The LFSMD is fortified and analyzed identically to the
        LFSM. The LFSMD is used to assess method precision when the observed concentrations of
        method analytes are low.

-------
   4.11. LABORATORY METHOD BLANK (LMB) - A blank matrix that is treated exactly the same as
        a sample including exposure to all glassware, equipment, solvents and reagents and surrogate
        standards that are used in the analysis batch. The LMB is used to determine if method analytes or
        other interferences are present in the laboratory environment, the reagents, or the apparatus.

   4.12. MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by vendors
        concerning a chemical's toxicity, health hazards, physical properties, fire, and reactivity data
        including storage, spill, and handling precautions.

   4.13. MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be reported as
        a quantitated value for a method analyte in a sample following analysis. This defined
        concentration can be no lower than the concentration of the lowest calibration standard for that
        analyte and can be used only if acceptable QC criteria for this standard are met.

   4.14. PRECURSOR ION - For the purpose of this method, the precursor ion is the protonated
        molecule ([M+H]+) or adduct ion of the method analyte. In MS/MS, the precursor ion is mass-
        selected and fragmented by collisionally induced dissociation (CID) to produce distinctive
        product ions of lower mass.

   4.15. PRODUCT ION - For the purpose of this method, a product ion is one of the fragment ions
        produced in MS/MS by CID of the precursor ion.

   4.16. SURROGATE STANDARD (SS) - A pure chemical(s) added to a standard solution in a known
        amount(s) and used to measure the relative response of other  method analytes that are
        components of the same solution. The surrogate standard must be a chemical that is structurally
        similar to the method analytes, has no potential to be present in samples, and is  not a method
        analyte.

   4.17. STOCK STANDARD SOLUTION (SSS) -  A concentrated solution containing one or more
        method analytes prepared in the laboratory using assayed reference materials or purchased from
        a reputable commercial source.
5.   INTERFERENCES

    Procedural interferences can be caused by contaminants in solvents, reagents, glassware  and other
    apparatus that lead to discrete artifacts or elevated baselines in the selected ion current profiles. All
    of these  materials  must routinely be  demonstrated to be free from interferences by  analyzing
    Laboratory Method Blanks (LMBs) (Section 10.4.1) under the same conditions as the samples (5).
    Subtraction of blank values from sample results is not performed.

    5.1. All reagents and solvents should be of pesticide grade purity or higher to minimize interference
        problems.  All glassware should be cleaned and demonstrated to be free from interferences.

    5.2. Matrix interferences may be caused by contaminants from the sample matrix, sampling devices
        or storage containers.  The extent of matrix interferences will vary  considerably from sample
        source  to sample  source, depending upon variations  in the  sample  matrix. Wipe  matrix
        interferences and contaminants are likely to be present and may have  an effect on the recoveries
        for the analytical procedure.  These interferences lead to elevated baselines and artifacts that may
        be interpreted as positives. Wipes were not pre-cleaned but were analyzed to ensure  that there
        were no interferences present. Any wipe materials containing interferences with the analytes of

-------
        interest were not used.

    5.3. Matrix  effects  are  known  phenomena  of ESI-MS  techniques,  especially  for coeluting
        compounds.  Managing the unpredictable suppression and enhancement caused by these effects
        is recognized as an integral part of the performance and verification of an ESI-MS procedure.
        The data presented in this procedure were designed to demonstrate that the procedure is capable
        of functioning  with  realistic  samples.   Each analyst  is encouraged to observe appropriate
        precautions and follow the  described QC procedures to help minimize the influence of ESI-MS
        matrix effects on the data reported.  Matrix effects include ion suppression/enhancement, high
        background and improper ion ratios.

6.   HEALTH AND SAFETY

    The toxicity and carcinogenicity of each reagent used in this method have not been defined precisely.
    However, each chemical compound was  treated as a health hazard. Exposure to these chemicals
    should be reduced to the lowest possible level and proper protective equipment should be worn for
    skin, eyes, etc.  Each laboratory is  responsible for maintaining an awareness of Occupational Safety
    and Health Administration (OSHA) regulations regarding the safe handling of chemicals used in this
    method.  A reference file of MSDSs that address the  safe handling of the chemicals should be made
    available  to  all personnel involved in the  chemical analyses or  subject to  potential exposure.
    Additional references are available (6-9).

7.   EQUIPMENT AND SUPPLIES

    References to specific brands of equipment and catalog numbers are provided solely as examples and
    do not constitute an endorsement of the use  of such products or suppliers. Materials tested for the
    wipe analysis of nerve agent degradation products are  described in Table 1.

      7.1 LC-MS/MS APPARATUS

           7.1.1    LIQUID CHROMATOGRAPHY (LC) SYSTEM - An analytical system complete
                   with a temperature programmable  liquid  chromatograph with  a  solvent mixer
                   (Waters, Milford,  MA - Acquity™  or equivalent able to perform the analyses as
                   described) and all required  accessories including syringes, solvent degasser, and
                   autosampler.

           7.1.2   ANALYTICAL COLUMN - Atlantis® - dC18, 100 mm x 2.1 mm, 3 |^m particle size
                  (Waters, Milford, MA, Catalog # 186001299), or equivalent.

           7.1.3   TANDEM MASS SPECTROMETER (MS/MS) SYSTEM - An MS/MS instrument
                  (Waters TQD™ or  similar instrument) can be used for analysis of the target analytes.
                  A mass spectrometer capable of MRM analysis with the capability to obtain at least
                  10 scans over a peak with adequate sensitivity is required.

           7.1.4   DATA SYSTEM - Waters' MassLynx™ software  (or similar software) interfaced to
                  the LC/MS that allows the continuous acquisition and  storage on machine-readable
                  media of all mass spectra obtained throughout the duration of the chromatographic
                  program.  Waters'  QuanLynx™  (or similar  software)  is used for all quantitative
                  analysis for data generated from the LC-MS unit.

      7.2 EXTRACTION DEVICE

-------
          7.2.1   SONICATOR (Fisher Scientific Catalog # 15-335-112) or equivalent.

     7.3 GLASSWARE AND MISCELLANEOUS SUPPLIES

          7.3.1   AUTOSAMPLER VIALS - Amber 2-mL autosampler vials with pre-slit Teflon®-
                 lined screw tops (Waters Corp., Milford, MA), or equivalent.

          7.3.2   DISPOSABLE STERILE SYRINGES - 10.0 mL ± 1% accuracy BD Safery-Lok™
                 syringes (Catalog No. 14-829-32, Fisher Scientific, Pittsburgh, PA), or equivalent.

          7.3.3   AUTO PIPETTES - 10.0 mL, 1000 uL, 100 uL and 10 uL ± 1% accuracy.

          7.3.4   DESOLVATION GAS  - Nitrogen gas generator or equivalent nitrogen gas supply.
                 Aids in the generation  of an aerosol of the  ESI liquid spray and should meet or
                 exceed instrument manufacturer's specifications.

          7.3.5   COLLISION GAS - Argon gas used in the collision cell in MS/MS instruments and
                 should meet or exceed instrument manufacturer's specifications.

          7.3.6   ANALYTICAL BALANCE - accurate to 0.1 mg; reference weights traceable to
                 Class S or S-l weights.

          7.3.7   National Institute of Standards and Technology (NIST)-traceable thermometer.

          7.3.8   STANDARD SOLUTION FLASKS  - Class A volumetric glassware

          7.3.9   SYRINGE FILTER - Millex® GV Syringe-driven polyvinylidene fluoride (PVDF) 13
                 mm  filter unit ,  0.22 jam (Millipore Corporation,  Billerica,  MA,  Catalog  #
                 SLGV013NL).

          7.3.10  WIPES - Dukal™, 2" x  2" - 12-ply sterile cotton gauze pads, individually packaged
                 (Fisher Scientific, Pittsburgh, PA, Catalog # 17986468).

          7.3.11  SAMPLE COLLECTION CONTAINERS -  Clean  125 ml Nalgene  polypropylene
                 straight-side jars with screw caps (Fisher Scientific, Pittsburgh, PA,  Catalog #11-
                 815-1OC), or equivalent.

          7.3.12  SAMPLE CONCENTRATION CONTAINERS - Sterile 15 mL conical graduated
                 plastic centrifuge tubes  (Fisher Scientific, Pittsburgh, PA, Catalog # 05-5 3 8-5 9A), or
                 equivalent.
8  REAGENTS AND STANDARDS

     8.1       REAGENTS AND STANDARDS

              When compound purity is assayed to be 98% or greater, the weight may be used without
              correction to  calculate the  concentration of the stock  standard.  Expiration times for
                                            7

-------
       prepared solutions  are suggested below, but laboratories should follow  standard QC
       procedures to determine when the standards should be replaced.  Label all standards and
       verify the correct grade of solvents.  Traceability of standards is  established by the
       manufacturer's specifications provided at time of purchase.

    8.1.1   SOLVENTS, REAGENTS and GASES - Acetonitrile (CAS # 75-05-8), Methanol
           (CAS  # 67-56-1),  and LC-MS  grade Water  (CAS  # 7732-18-5),  HPLC  mass
           spectrometry pesticide grade or equivalent, demonstrated to be free of analytes and
           interferences. Formic Acid (Chemical Abstracts Service (CAS) # 64-18-6). Nitrogen
           is used for the generation of aerosol of the ESI liquid spray, and purity should meet
           instrument manufacturer's specifications.  Argon  is used as the collision gas in
           MS/MS   applications,   and  purity  should  meet  instrument   manufacturer's
           specifications.

    8.1.2   MOBILE PHASE A - Solution A  consisted of LC-MS grade water  and 0.2% of
           formic acid to prevent microbial growth. To prepare 0.5 L, add 1 mL of formic acid
           and dilute to 0.5 L mark with water. This solvent system is prone to some microbial
           growth and should be replaced at least once a week.

    8.1.3   MOBILE PHASE B- Solution B  was comprised of acetonitrile and 0.2% of formic
           acid. To prepare 0.5 L, add 1 mL  of formic acid and dilute to 0.5  L mark with
           acetonitrile.

    8. 1 .4   TARGET ANALYTES - MPA (Catalog #: 289868) and EMPA (Catalog #: 1 12062)
           were purchased from Sigma-Aldrich (St. Louis , MO).  IMPA (Catalog #: ERI-015),
           DIMP (Catalog #: ERD-083), PMPA (Catalog #: ERP-083), and EHDMAP (Catalog
           #: ULM-6091-1.2) were purchased from Cerilliant (Round Rock, TX).

    8.1.5   SURROGATE  ANALYTES  - MPA-d3  (Catalog  #: DLM-6196-1.2), PMPA-13C6
           (Catalog #: CLM-6620-1.2)and DIMP-d14 (Catalog #: ERD-086) were purchased
           from Cerilliant.

8.2    STANDARD SOLUTIONS

       When compound purity is assayed to be at least 98% or greater, the weight can be used
       without correction to calculate the concentration of the  stock standard.  Stock standards
       and all subsequent  solutions should be replaced when  analyzed solution concentrations
       deviate more than ± 20% from the prepared concentration. Standards are stored protected
       from light (amber flasks) and at 4 °C (± 2 °C).  Standards are estimated to be stable for at
       least a month as long as water is not present.  Although stability times are suggested,
       laboratories should utilize QC practices to determine when standards should be replaced.

    8.2.1       SURROGATE STOCK  STANDARD  SOLUTION (Surrogate  SSS) (10-1000
           A standard solution may be prepared from certified commercially available methanol
                                                                                    •6
solutions or neat compounds. Isotopically-labeled surrogates (MPA-d3, PMPA- C
and DIMP-di4) were purchased as methanol solutions. The surrogate is added to a 10
mL volumetric flask to achieve  a concentration of approximately ten times the
highest calibration concentration (ten times calibration 7) in solution (i.e., 900  |oL of
                                      8

-------
              MPA-d3 and PMPA-13C6 and 18 nL of DIMP-d14 were added to a 10 mL volumetric
              flask and diluted to the mark with methanol). Surrogate stock standard solutions are
              stable for at least a month when stored at 4 °C.

              (NOTE: Although the listed analytes were used as surrogates in this method, they
              could also be used as internal standards for quantitation purposes. However, further
              evaluation would be necessary to ensure that they are viable internal  standards and
              meet QC requirements.)

       8.2.2   ANALYTE STOCK STANDARD SOLUTION (AS)

              Standard solutions may be prepared  from certified commercially available neat
              compounds.    MPA  and BMP A  were  purchased  as  a  neat  solid and  liquid,
              respectively.   Separate methanol  solutions  (1000 |o,g/mL) containing MPA and
              EMPA were used to make the analyte  stock standard solution.  DIMP, EHDMAP,
              IMPA, and  PMPA were purchased as methanol  solutions. A standard methanol
              solution with a concentration of 3  |o,g/mL (ppm) was made in a 25 mL volumetric
              flask containing DIMP, EMPA, MPA, PMPA, and IMPA (i.e., 18 jiL of DIMP, 30
              HL of EMPA, 75 jiL of MPA, 18 jiL of PMPA and 90 jiL of IMPA are each added to
              a 25 mL volumetric flask and diluted to the mark with methanol).  EHDMAP is not
              added to the initial stock standard solution because it is not stable over the suggested
                                                                                  OO
              stability period when added to the  methanol solution. EHDMAP and the surrogate
              analytes are added to calibration standard solutions only when the solutions are ready
              for use. The calibration standards and spike solutions are made from the appropriate
              dilution of this analyte  stock standard.  The analyte stock standard solution is stable
              for at least a month when stored at 4 °C.

       8.2.3   CALIBRATION STANDARD SOLUTION (CAL)

              Dilution of the 3 jog/mL methanol solution can be used to obtain a 750 ng/mL (ppb)
              solution in water.  A calibration stock standard solution (Level 7) is prepared from
              the Analyte Stock Standard Solution (AS) and SSS by adding, 2.5 mL of AS, 9 joL of
              EHDMAP, and 1 mL of the SSS (i.e., 2.5 mL of the AS containing DIMP, EMPA,
              MPA, PMPA, and IMPA, 9 nL of EHDMAP, and 1 mL of the SSS are added to a 10
              mL volumetric flask and diluted to the mark with LC-MS grade water). From Level
              7, further dilutions are performed with LC-MS grade water to prepare Levels 6
              through 1, as shown in Table 2.
SAMPLE COLLECTION. PRESERVATION AND STORAGE

  9.1       SAMPLE COLLECTION

       9.1.1   Volatile organic analysis (VOA) vials and Nalgene  containers were both used for
              sample collection and both were deemed adequate for use, but Nalgene containers
              were specifically used in this method. Other vessels may be used as long as they are
              tested and verified to ensure they do not contain any interfering compounds.  As an
              example  for field  samples, the  field  samplers  would collect  samples with the
              appropriate water-wetted wipe  and place the wipes in ajar with a cap (e.g.,  125 mL
              Nalgene polypropylene  straight-sided jar with a polypropylene screw  cap) and ship
              the jar containing the sample  to the laboratory.   The Nalgene containers  did not
              present contamination problems nor did the results suggest that  the analytes  of

-------
                  interest adhere to the jars, so Nalgene containers can be used instead of glass VOA
                  vials used in standard practice.

           9.1.2   The wipe is wetted with 1 mL of LC-MS grade water, sufficient to wet the wipe.  The
                  surface is wiped in a Z-like pattern horizontally across a defined surface (100 cm2)
                  (Attachment 19.3), folded, then used to  wipe the  same surface in a Z-like pattern
                  vertically across a defined surface (100  cm2).  The wipe is placed into a 125 mL
                  Nalgene polypropylene straight-sided jar with a polypropylene screw cap. Surrogates
                  (66.6 joL of the  SSS) and LC-MS grade water (5  mL) are added to the jar. Field
                  and/or matrix  blanks are needed,  according  to  conventional sampling  practices;
                  therefore, one blank sample coupon was analyzed in every sample extraction batch.

     9.2 SAMPLE STORAGE AND HOLDING TIMES

           9.2.1   Wipe samples  should be extracted as soon as possible after collection  but must be
                  extracted within 30 days of collection.   Samples  not immediately analyzed from a
                  particular site should  be carefully characterized to  ensure there is no interaction with
                  the wipe or a specific surface to  cause interferences or degradation of the analytes.
                  An LFSM  can  be generated for the appropriate time  period to verify  such an
                  occurrence. Samples can be stored up to 30 days (Table 2) at 4 °C (± 2 °C).

10  QUALITY CONTROL

        10.1   QC requirements include  the performance of an initial demonstration of capability (IDC)
            and ongoing QC requirements that must  be met to generate data of acceptable quality when
            preparing and analyzing samples.  This section describes the QC parameters, their required
            frequencies and performance criteria.  A precision  and accuracy study (P&A,  as shown in
            section 19.2) as well  as a Detection Limit (DL) study (Table 3  and section 19.1) must be
            performed to demonstrate laboratory capability. Laboratories are  encouraged to institute
            additional QC practices to meet their specific needs.

        10.2   INITIAL DEMONSTRATION OF CAPABILITY (IDC)

               The IDC must be  performed successfully  prior to the initiation of analysis of field
               samples. Prior to conducting an IDC, an acceptable Initial Calibration must be generated
               as outlined in Section 11.2.

           10.2.1  INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND

                  Any time a new lot of solvents,  reagents, filters and autosampler vials is used, the
                  LMB must be demonstrated to be  reasonably  free of contamination (i.e., that the
                  criteria are  met  as stipulated in  Section 10.4.1).   The LMB is used to ensure that
                  analytes  of interest   or other  interferences  are  not  present  in  the laboratory
                  environment, the solvent, or the apparatus.

                  NOTE: Good laboratory practices indicate the  use of a blank before  and after
                  analyzing a calibration curve for an instrument to ensure that no carryover will occur.
                  If the required criteria are not met and samples were not free of contamination, then
                  the source  of the contamination should be identified and eliminated before the
                  performance of any analysis.
                                              10

-------
           10.2.2  INITIAL DEMONSTRATION OF PRECISION AND ACCURACY (P&A)

NOTE: Because porosity of the wiped surface will inevitably have an effect on analyte recovery from the
surface, accuracy results between calculated values and true values may differ from surface to surface.
The precision and accuracy results are based on the wipe used on the laminate (Formica®, Formica
Corp., Cincinnati, OH) surface because (1) the laminate surface has been shown to be free of
contamination, (2) this surface results in minimal surface interaction between the chemical and the
surface, and (3) the laminate is a relatively nonporous surface.

                  For a  P&A,  prepare a check standard  containing DIMP, EMPA, MPA, PMPA,
                  EHDMAP, and IMPA near or below the midpoint concentration of the calibration
                  range.  This check standard  should be analyzed with a minimum of four replicates.
                  For this study, four different concentrations are chosen with seven samples each. The
                  check samples are analyzed according to Section 12.

           10.2.3  The average percent recovery (X), standard deviations  (a) and the percent relative
                  standard deviation (%RSD)  of the recoveries are calculated for each analyte.  The %
                  RPD limit of < 30% should be applied to all replicate analyses.

           10.2.4  MINIMUM REPORTING LEVEL (MRL)

                  Establish a target concentration for the MRL based on the intended use of the
                  method. Establish an Initial Calibration (Section 11.2). The lowest CAL standard
                  used to establish the initial calibration must be at or below the MRL concentration. If
                  the MRL concentration is too low, ongoing QC requirements may fail repeatedly,
                  and the MRL must be determined again at a higher concentration.  The MRL
                  reported in this study is the lowest calibration level. The MRL is validated following
                  the procedure below.

                10.2.4.1   Fortify, extract, and analyze seven replicate LFBs at the proposed MRL
                         concentration.  Calculate the mean measured concentration (Mean) and
                         standard deviation for these replicates.  Determine the Half Range for the
                         prediction interval of results (HRPIR) using the equation below

                                                  HRPIR = 3.963s

                         where
                            s      = the standard deviation
                            3.963  = a constant value for seven replicates (10).

                10.2.4.2   Confirm that the upper and lower limits for the Prediction Interval of Result
                         (PIR = Mean j_ HRPIR) meet the upper and lower recovery limits as shown
                         below

                         The Upper PIR Limit must be <150% recovery.

                                      Mean + HRP1R
                                  FortifiedConcentration

                         The Lower PIR Limit must be > 50% recovery.


                                              11

-------
                               Mean-HRP1R
                           FortifiedConcentration

    10.2.5  CALIBRATION VERIFICATION

           Mid-level and low-level samples from the calibration curve should be analyzed to
           confirm the accuracy of the  fit of the calibration curve/standards after the end of
           sample batches.

10.3   METHOD DETECTION LIMITS (MDL)

       The procedure for the determination of the laboratory detection and quantitation limits for
       the EPA approach follows 40 CFR Part 136, Appendix B.  MDLs represent the minimum
       concentration at which there is a high  degree of statistical confidence that, when the
       method reports that an analyte is present, that analyte is actually present (i.e., a low risk
       of false positives).

    10.3.1  DETERMINATION OF LABORATORY INSTRUMENT DETECTION LIMITS
           (IDLs)

           The laboratory IDL can be  used to establish an estimate of the initial spiking
           concentration  used for determination of the  MDL,  although other approaches for
           determining the initial spiking concentration may be used. The  laboratory IDL is
           determined for each analyte as a  concentration that produced an  average signal-to-
           noise (S/N) ratio in the range of 3:1 - 5:1 for at least three replicate injections.  For
           example,  successively lower concentrations of the analytes are injected until the S/N
           ratio  is in the  range of 3:1  -  5:1.   Replicates are then injected at that  target
           concentration to ensure that the average S/N of the replicates was within the 3:1 - 5:1
           range.   Note  that since  linearity  of  S/N  ratio  with  increasing or decreasing
           concentration cannot be assumed, the concentrations  determined  via this procedure
           are necessarily approximate.

    10.3.2  DETERMINATION OF LABORATORY METHOD DETECTION LIMIT (MDL)

           Method Detection Limits (MDLs)  represent the optimal detection achieved by  a
           laboratory in a matrix of interest.  The  analyte spiking solution, containing all six
           analytes, was added to the  surface (section 19.3).  The solution on the surface was
           allowed to completely dry and  wiped using  a wetted-cotton gauze  wipe.  Wipe
           extracts from the laminate  coupons  are used for the determination of the MDL for
           surface samples.   The 40 CFR Part  136,  Appendix  B  procedure is followed,
           particularly with regard to spike levels used. Replicate reference matrix samples are
           spiked at  a level between 1-5 times the estimated detection level (e.g., suggested by
           the IDL procedure in 9.3.1).  The resulting MDL must be within  10 times the spike
           level used, or the MDL determination would be repeated using a more appropriate
           spike level. Full method sample preparation procedures to prepare and analyze at
           least  seven replicates of the  spiked clean matrix of interest are  used.   Apply the
           following  equation to the analytical results (Student's t-factor is dependent on the
           number of replicates used; the value 3.14 assumes seven replicates):
                                      12

-------
                                                 MDL =!(„_!,!_„ = 0.99) XSD

          where

          MDL = method detection limit
          t           = Student's t value for the 99% confidence  level with n-1 degrees of
           (n-l,l-a = 0.99)
          freedom (for seven replicate determinations, the Student's t value is 3.143 at a 99%
          confidence level),
          n = number of replicates, and
          SD = standard deviation of replicate analyses.
          a = standard deviation of the percent recovery

          Data for MDLs are shown in Table 3 and Section 19.1.
10.4  ONGOING QUALITY CONTROL (QC) REQUIREMENTS

       10.4.1  LABORATORY FORTIFIED BLANK (LFB)

          An LFB is required with each extraction batch to confirm that potential background
          contaminants are not interfering with identification or quantitation of the target
          analytes.  If there is a contaminant within the retention time window preventing the
          determination of the target analyte, the source of the contamination should be
          determined and eliminated before processing samples. LFBs include cotton gauze
          wipes wetted with water.

       10.4.2  LABORATORY METHOD BLANK (LMB)

          An LMB  is prepared and analyzed with each extraction batch, using LC-MS grade
          reagent water, for confirmation that there are no background contaminants interfering
          with the identification or quantitation of the target analytes.  If there is a contaminant
          within the retention time window preventing the  determination of the target analyte,
          the  source  of  the contamination should  be determined  and  eliminated before
          processing samples.  LMBs include the  extracted  wipe  used  to wipe the  surface
          coupon.

       10.4.3  CONTINUING CALIBRATION CHECK (CCC)

          CCC standards are analyzed at the beginning and end of each  analysis batch.  The
          CCC is analyzed periodically to verify the accuracy of the existing calibration for
          analytes near the  midpoint of the calibration range and/or near the MRL.  CCC
          values  should  be  specified by  the sample submitter's Data  Quality Objectives
          (DQOs) or fulfill other QC requirements, such as LFSM acceptance).

       10.4.4  LABORATORY FORTIFIED SAMPLE MATRIX (LFSM)

          A LFSM is analyzed to determine that spike accuracy for a  particular sample matrix
          is not  adversely affected by  chemical interactions between  target analytes  and
          experimental matrices (i.e., coupon/wipe materials). If a variety of sample matrices
          are analyzed, performance should be established for each surface.
                                      13

-------
10.4.4.1Within each analysis batch, an LFSM is prepared and analyzed at a frequency of one
       sample matrix for every twenty samples. The LFSM is prepared by spiking a sample
       with the appropriate amount of AS (Section 8.2.2). Select a spiking concentration that
       is greater than or equal to the matrix background concentration, if known.  Records
       are maintained of the surface target compound spike analyses, and the average percent
       recovery (X)  and the standard  deviation of the percent recovery (a) are calculated.
       Analyte recoveries  may exhibit bias  for certain matrices.  Acceptable recoveries are
       50-150% if a low-level concentration  near or at the MRL (within a factor of 3) is used.
       If the recovery does not fall within this range, check with a CCC or prepare a fresh AS
       solution for analysis. If the  recovery of any analyte  still falls outside the designated
       range and the laboratory performance for that analyte is shown to be in control in the
       CCCs,  the recovery is judged to be matrix biased. The result for that analyte in the
       unfortified sample is labeled suspect/matrix to inform the data user that the results are
       suspect due to matrix effects.

    10.4.5  SURROGATE STANDARD

        All samples  (CCCs, LFBs, LMBs, LFSMs, LFSMDs, FDs, and CAL standards) are
        spiked with  surrogate standard spiking solution as  described in Section 8.2.1.  An
        average percent recovery of the surrogate compound and the standard  deviation of
        the percent recovery (REC) are calculated and updated regularly.

    10.4.6  FIELD DUPLICATE (FD) OR LABORATORY FORTIFIED SAMPLE
           MATRIX DUPLICATE (LFSMD)

        Within each  analysis batch, a minimum of one FD or LFSMD should be  analyzed for
        every  twenty samples.  Target compound spike accuracy  in the  sample matrix is
        monitored and updated regularly.   Duplicates check the precision associated with
        sample collection, storage  and laboratory procedures.  Records are  maintained of
        spiked matrix  analyses and the average  percent recovery (X) and  corresponding
        standard deviation (a) are calculated. FD/LFSMD samples must be incorporated into
        the field  sampling  plan.   If the  laboratory  did not receive  FD samples  for
        determination of site-specific P&A,  the laboratory will evaluate the site data quality
        based  on the LFSM data, if there is  sufficient sample in the site samples to conduct
        an  analysis.  FD/LFSMD recovery results will be used  for site-specific P&A data.
        LFSM data are used as FD/LFSMD sample data for this  study.  RPD values  should
        be  < 30% for FD/LFSMD samples.

      10.4.6.1  Calculate the relative percent difference (RPD) for duplicate  measurements
              (FDj and FD2) using the equation:

                                FD, -FD9
                       RPD = 7
                                   +FD2)/2

            RPDs for Field Duplicates should be < 30% for each analyte. Greater variability
            may be observed when Field Duplicates have analyte concentrations at or near
            the MRL (within a factor  of two times  the  MRL concentration).  At these
            concentrations, FDs must have RPDs that are < 50%. If the RPD of an analyte
            falls outside the designated range and the laboratory performance for the analyte
                                   14

-------
                      is shown to be in control in the CCC and in the LFB, the precision is judged
                      matrix influenced.  Report the result for the  corresponding analyte in  the
                      unfortified sample as "suspect/matrix."

                10.4.6.2 If an LFSMD is analyzed instead of an FD, calculate the RPD for the LFSM
                       and
                                            LFSM-LFSMDl
                       using   the  RPD =
                       equation:           (LFSM + LFSMD)/2
                      RPDs for duplicate  LFSMs should be  < 30% for each  analyte.   Greater
                      variability  may be  observed  when  the  matrix is  fortified  at  analyte
                      concentrations at or  near the MRL  (within a factor of two times the  MRL
                      concentration).  LFSMs at these concentrations must have RPDs that are < 50%.
                      If the RPD of an analyte  falls outside the designated range and the laboratory
                      performance for the analyte is shown to  be in control in the CCC and in the
                      LFB, the  precision is judged matrix influenced.  Report the result for the
                      corresponding analyte in the unfortified sample as "suspect/matrix."
11   INSTRUMENT CALIBRATION AND STANDARDIZATION

   All laboratory equipment should be calibrated according to manufacturer's protocols.  Demonstration
   and documentation of acceptable mass spectrometer (MS) tuning and initial calibration is necessary
   prior to sample analysis. Verification of the tuning of the MS must be repeated each time instrument
   modification/maintenance is performed and prior to analyte calibration.  After initial calibration is
   successful, a CCC ( at the appropriate concentration described in section 10.4.2) should be performed
   at the beginning and end of each analysis batch.

       11.1 CALIBRATION OF MASS SPECTROMETER

              Calibrate the mass scale of the  mass spectrometer as prescribed by the manufacturer.
              The  mass  calibration file is saved in the  mass spectrometer software file folder
              (MassLynx™ or similar software).  The  mass calibration solution used in this method is a
              mixture of NaCsI provided by the manufacturer.  Other calibration solutions can also be
              used per instrument manufacturer's specifications.

       11.2 INITIAL CALIBRATION FOR ANALYTES

               11.2.1  ESI negative mode is the preferred choice for this method  due  to the optimal
                     conditions and advantages (e.g., greater  peak intensity, few interferences, and
                     lower background) in ESI  negative mode over ESI positive  mode.  However,
                     ESI positive mode may be used  if matrix interferences become problematic. The
                     data are presented in  both modes in some tables, but for clarification, only ESI
                     negative mode will be discussed in the method.

               11.2.2  Optimize the [M-H]~  ion in ESI negative mode for each analyte  by infusing an
                     appropriate calibration solution at a flow rate similar to that the flow rate used for


                                             15

-------
                      the LC  separation.  Adjust MS parameters (voltages, temperatures, gas flows,
                      etc.) until optimal analyte responses are achieved.  Optimize the product ion by
                      following the same procedures as for the [M-H]~ ion.  Ensure that there are at
                      least 10 scans  across the peak  for  optimal precision.   ESI-MS and MS/MS
                      parameters utilized during development of this method are presented in Tables 4a
                      and 4b and 5.

               11.2.3  Establish LC operating conditions that will optimize peak resolution and shape.
                      Suggested LC conditions  (listed in  Table 6) may not  be  optimal for all  LC
                      systems.

               11.2.4  The  initial   calibration  contains  a seven-point  curve  using   the  analyte
                      concentrations prepared in section 8.2.3  and shown in Table 7.  The  lowest
                      calibration curve standard must be at the MRL.  The calibration curve and all
                      samples should be analyzed in a low to high concentration regimen  so carryover
                      is  less of a  concern in case the LC  cleaning cycle does not clean the system
                      adequately between injections.  Verify that all analytes have been properly
                      identified and  quantified  using software programs.   Integrate  manually, if
                      necessary, in accordance with laboratory quality assurance plans  Depending on
                      the instrument, sensitivity  and calibration  curve responses may vary.   At a
                      minimum, a five-point linear or a six-point quadratic calibration curve will be
                      utilized  for all analytes. If the polynomial type excludes the point of origin, use a
                      fit weighting of 1/X to give more weighting to the lower concentrations.  The
                      coefficient of determination (r2) of the linear fit should be greater than or equal to
                      0.98.  If one of the calibration standards other than the high or low standard
                      causes the r2 to be <0.98, this point must be re-injected or  a new calibration
                      curve must be analyzed.  If the  low  and/or high point is excluded, a six-point
                      curve is acceptable but  the calibration  range and  reporting limits  must be
                      modified to reflect this change.  The  r2 of the quadratic curve should be greater
                      than or  equal to 0.99. If one of the calibration standards other than the high or
                      low standards causes the r2 to be <0.99, follow the same procedure  given above
                      for a linear fit.  A calibration curve and an instrument blank will be analyzed at
                      the beginning of each batch or daily to ensure instrument stability (9).  When
                      quantitated, each calibration point for each analyte should calculate  to be within
                      70-130% of its true value. The lowest  CAL standard should calculate to be within
                      50-150% of its true  value.  A new curve will be generated daily. The calibration
                      method  is used to quantify all samples.

       11.3 QUANTITATION OF ANALYTES

               The quantitation of the target analytes is accomplished with  quantitation software as it
               relates  to  each  specific instrument (9).  An external calibration  is used along with
               monitoring MPA-d3, PMPA-13C6 and DIMP-di4 surrogate  recoveries.  Refer  to Tables 4a
               and 4b for the MRM transitions and retention times utilized during  the development of
               this method.
12   ANALYTICAL PROCEDURE
       12.1 SAMPLE PREPARATION
                                               16

-------
       12.1.1   Samples were collected and stored as described in Section 9. Surrogates (MPA-
               d3, PMPA-13C6 and DIMP-d14) are added first, then LC-MS grade water (5 mL) is
               added to the jar.  Sonicate each jar containing the solution for approximately 15
               minutes in a water bath at room temperature with no heat required.

       12.1.2   After sonication,  decant the extraction solvent into a 10 cc lock-tip sterile fitted
                                   ®
               syringe with a Millex  GV syringe driven filter unit, PVDF  filter (0.22  jam),
               transferring the  filtered sample  to a  sterile 15-mL  polypropylene tube  (or
               equivalent).

       12.1.3   Transfer (via pipette) to a standard 2 mL sample vial.

           NOTE:  Calibration  standards are not  filtered through the syringe-driven filter units
           because  no particulates are present. The filters and syringes used in this study were
           not shown to affect analyte concentrations.  If alternate filtering is  incorporated, the
           filters should be  subjected to QC requirements to ensure they do not introduce
           interferences or retain the target analytes.

12.2   SAMPLE ANALYSIS/ANALYTICAL SEQUENCE

       12.2.1   Use  the same Liquid Chromatography/Mass Spectrometry conditions established
               per guidance described in Section 11 and summarized in Tables 4a, 4b, 5 and 6.

       12.2.2   Prepare an  analytical batch that includes all QC samples and  surface samples.
               The  first  sample  to be analyzed is a 10 uL injection of a blank (LC-MS grade
               reagent water) on column followed by the calibration curve.

       12.2.3   Update the  calibration file and print a calibration report.  Review the report  for
               calibration outliers and make area corrections by manual integration, if necessary
               and  appropriate.   If corrections have  been made,  update the  calibration file,
               noting the changes, and regenerate  a calibration report.  Alternatively, re-analyze
               "nonconforming" calibration level(s) and repeat the above procedures.

       12.2.4   The  first sample analyzed after the calibration  curve is  an additional blank (LC-
               MS  grade reagent water) to ensure there  is no carryover (11).  If the initial
               calibration data are acceptable, begin analyzing samples, including QC and blank
               samples, at their  appropriate frequency injecting the  same size aliquots (10 joL)
               under the same conditions used to analyze CAL standards.  The ending  CCC
               must have  each  analyte  concentration within 30% of the  calculated true
               concentration or  the affected analytes  from that  run  must be  qualified  as
               estimates or the samples must be re-analyzed with passing criteria to remove the
               qualification.

       12.2.5   If the absolute amount  of a target  compound exceeds the working  range of the
               LC-MS system (see Level 7 in Table 7),  the prepared sample is  diluted with
               water and re-analyzed along with additional samples that may have  run after the
               sample known to exceed the calibration range, because of the possibility of
               carryover. Care must be taken to ensure that there is no carryover of the analyte
               that  has  exceeded the calibration range.  If the amount of analyte exceeds the
               calibration range, a blank sample should be analyzed afterward to demonstrate no
               carryover will occur.
                                       17

-------
               12.2.6 At the conclusion of the data acquisition, use the same software that is used in the
                     calibration procedure  to  identify  peaks of interest from  the  predetermined
                     retention time windows. Use the data software to examine the ion abundances of
                     the peaks in the chromatogram to  identify  and compare retention times in the
                     sample chromatogram with the retention time of the corresponding analyte peak
                     in an analyte standard.
13   DATA ANALYSIS AND CALCULATIONS

       13.1 QUALITATIVE AND QUANTITATIVE ANALYSIS

               13.1.1  Complete chromatographic resolution is  not  needed for accurate and precise
                      measurements of  analyte concentrations when using MS/MS.  An external
                      calibration is used when monitoring the MRM  transitions of each analyte.
                      Quantitation software is utilized to conduct the quantitation of the target analytes
                      and surrogate standards.   The  MRM transitions of each analyte are used for
                      quantitation and confirmation.  The MRM transition serves as a confirmation by
                      isolating the precursor ion, fragmenting the precursor ion to the product ion, and
                      relating the transition to the retention time in the calibration standard (9).

               13.1.2  Computer  programs used  for  analysis  of data include instrumentation  and
                      quantitation software.  Manual integration may be necessary for some peak areas
                      if the peak area is not integrated properly (i.e.,  the integration for the peak is not
                      fully performed by the  instrument's software, which will be noticeable by visual
                      inspection  of each  peak).  Inspect all integrated peaks  for visible  integration
                      errors and manually integrate as necessary to ensure consistent integration of
                      other peaks and/or known calibration peaks. Any manual integration should be
                      carried out  by a qualified analyst, noted, and checked against quality control
                      procedures (sections 10 and 11.3).

       13.2 Prior to reporting  data, the  chromatogram  should  be reviewed for any incorrect  peak
            identifications.  The retention time window of the MRM transitions must be  within 5% of
            the retention time of the analyte standard.  If this is not true, the calibration curve needs to
            be re-analyzed to see  if there was  a shift in retention times during the  analysis and the
            sample needs to be re-injected.  If the retention time  is still incorrect in the sample, the
            analyte is referred to as an unknown. If peaks need to be manually adjusted due to incorrect
            integration by the program, clarification of where professional judgment was used to  alter
            the peaks should be documented during the data reduction and verification process.
14  METHOD PERFORMANCE

       14.1 PRECISION, ACCURACY AND DETECTION LIMITS

              14.1.1  Tables for precision, accuracy and detection limit results for a single laboratory
                      study are presented in Sections 19.1 and 19.2 and Table 3.

       14.2 RECOVERIES AND PRECISION FOR OTHER SURFACE TYPES
                                              18

-------
               14.2.1  Section 19.2 lists recoveries and precision of target analytes for a variety of other
                      surfaces.
        14.3 WIPE STORAGE STABILITY STUDY

               14.3.1  Extract storage was conducted on the laminate surface fortified with the targeted
                      method analytes. Precision and accuracy (n = 4) of the extracts were analyzed on
                      days 0, 2, 3, 7, 14, 21, and 30 days and are reported in Table 2.

        14.4   PROBLEM ANALYTES AND SURFACES

            14.4.1  TARGET ANALYTES ON UNCLEANED SURFACES

                   DIMP and some  EHDMAP  recoveries may be problematic due to the volatility or
                   rapid decomposition of these specific compounds (12).  Analysts should be aware
                   that these  two specific compounds may  not  be present within the tested sample
                   matrix and plan accordingly. EHDMAP detection limits  in this method are based on
                   samples  extracted within  the  same day.   Due to the  degradation of EHDMAP,
                   samples  analyzed after 24 hours may reflect different results. Furthermore, ESI (+)
                   analysis  results are problematic for certain compounds (e.g., IMPA and MPA) due to
                   possible  electrospray enhancement/suppression effects, whereas ESI (-)  results tend
                   to be more reliable. Both ESI (+) and (-)  results are presented. However, detection
                   limits are based on ESI (-) data. Wood surfaces resulted in poor recoveries outside
                   the range of this procedure.  As a result, the method should not be used to identify
                   these analytes on  a wood surface.  Although porosity of the surface is most likely the
                   culprit for low  recoveries,  further analysis  should be  performed to determine
                   definitive  reasons for poor recoveries from  the  surface. Direct extraction  of the
                   analytes  from wood could be used to elucidate whether or not chemical interactions
                   are occurring between target analytes and compounds found in a wood matrix.
15   POLLUTION PREVENTION

       15.1  This method utilizes small volumes of organic solvent and small quantities of pure analytes,
            thereby minimizing the potential hazards to both analyst and environment.

       15.2  For information about pollution prevention that may be applicable to laboratory operations,
            consult "Less is Better: Laboratory Chemical Management for Waste Reduction" available
            from the American Chemical Society's Department of Government Relations and Science
            Policy,   1155   16th   Street  N.W.,  Washington,   D.C.,   20036   or  on-line  at
            http://www.acs.org/content/dam/acsorg/about/governance/committees/chemicalsafety/public
            ations/less-is-better.pdf (accessed August 15, 2013).

 16  WASTE MANAGEMENT
        16.1  The analytical procedures described in this procedure generate relatively small amounts of
             waste since only small amounts of reagents and solvents are used. Laboratory waste
             management practices must be conducted consistent with all applicable rules and
             regulations, and laboratories should protect the air, water, and land by minimizing and
                                              19

-------
     controlling all releases from fume hoods and bench operations. Also, compliance with any
     sewage discharge permits and regulations is required, particularly the hazardous waste
     identification rules and land disposal restrictions.

16.2  Each laboratory should determine with federal and local officials how to safely dispose of
      field and QC samples. Waste containers should be properly labeled to identify the
      contents. Remember to attach the appropriate chemical waste label, date the beginning of
      collection before using the container and follow all appropriate federal and local waste
      disposal requirements.
                                       20

-------
17  REFERENCES
       1.  U.S.  Environmental  Protection Agency (EPA),  2012. Selected Analytical Methods for
           Environmental Restoration Following Homeland Security Events (SAM). EPA/600/R-12/555
           July  2012.  Cincinnati,  Ohio:  United  States Environmental  Protection Agency,  Office of
           Research and Development, National Homeland Security Research Center.

       2.  Code of Federal Regulations, 40 CFR Part 136, Appendix B. Definition and Procedure for the
           Determination of the Method Detection Limit - Revision 1.11.

       3.  Federal Advisory Committee on Detection and Quantitation Approaches and Uses in Clean
           Water Act Programs. Final Report. (Submitted to U.S. Environmental Protection Agency.)
           December 28, 2007.

       4.  Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, W.L. Budde, "Trace Analyses for
           Wastewaters." Environ. Sci. Technol. 1981, 15, 1426-1435.

       5.  Standard Practices for Preparation of Sample Containers  and for Preservation of Organic
           Constituents,  ASTM Annual Book of Standards, Part 31, D3694-78. Philadelphia:  American
           Society for Testing and Materials.

       6.  OSHA Safety and Health Standards, General Industry (29CRF1910). Occupational Safety and
           Health Administration, OSHA 2206 (Revised, July 1, 2001).

       7.  Carcinogens-Working with Carcinogens, Publication No. 77-206. Atlanta, Georgia:
           Department of Health, Education, and Welfare, Public Health Service, Center for Disease
           Control, National Institute of Occupational Safety and Health, August 1977.

       8.  Safety  in  Academic Chemistry Laboratories, American Chemical  Society Publication,
           Committee on Chemical Safety, 7th Edition.

       9.  "Prudent  Practices in the Laboratory:  Handling and Disposal of Chemicals," National
           Academies Press (1995), available at http://www.nap.edu (accessed August 15, 2013).

       10. Winslow, S.D.,  Pepich, B.V., Martin, J.J., Hallberg, G.R, Munch, D.J., Frebis, C.P., Hedrick,
           E.J.,  Krop,  R.A. "Statistical Procedures  for Determination and Verification of  Minimum
           Reporting Levels for Drinking Water Methods." Environ. Sci. Technol. 2006, 40, 281-288.

       11. Peters, F.T.; Drummer,  O.H.; Musshoff,  F. "Validation of New Methods", Forensic  Science
           International 2007, 165 (2): 216-224

       12. ASTM Standard D7597 - 09el, 2009 "Standard Test Method for Determination of Diisopropyl
           Methylphosphonate, Ethyl Hydrogen Dimethylamidophosphate, Ethyl Methylphosphonic
           Acid, Isopropyl Methylphosphonic Acid, Methylphosphonic Acid and Pinacolyl
           Methylphosphonic Acid in Water by Liquid, Chromatography/Tandem Mass Spectrometry"
           West Conshohocken, PA: ASTM International. 2009, DOI: 10.1520/D7597-09E01 ,
           www.astm.org.
                                              21

-------
18  TABLES AND VALIDATION DATA
 Item          Title
 Table 1. Materials Tested for the Wipe Analysis of Nerve Agent Degradation Products	2
 Table 2.  Holding Time Sample Stability of Nerve Agent Degradation Analytes of Wipe Samples in ESI Negative Mode	2
 Table 3.  Method Parameters for Nerve Agent Degradation Products	3
 Table 4a. ESI (+) MRM Ion Transitions, Retention Time (RT) and Variable Mass Spectrometer Parameters	4
 Table 4b.  ESI (-) MRM Ion Transitions, Retention Time (RT) and Variable Mass Spectrometer Parameters	4
 Table 5. ESI (+) and (-) MS/MS Conditions	5
 Table 6. Liquid Chromatography Gradient Conditions	5
 Table 7.  Target Concentrations of Calibration Standards Used During the Development of this Method (ng/mL)	6

-------
                         Table 1. Materials Tested for the Wipe Analysis of Nerve Agent Degradation Products
Material
Glass
Vinyl Tile
Laminate
Wood (southern pine, pre-treated)
Galvanized steel
Painted Drywall (BEHR® latex paint)
Manufacturer/Vendor
Carolina Glass Co./Lowe's
Armstrong/Home Depot
Wilsonart Laminate/Home Depot
Home Depot
McMaster-Carr
BEHR/Home Depot
           Table 2. Holding Time Sample Stability of Nerve Agent Degradation Analytes of Wipe Samples in ESI Negative Mode
ESI (-) Mode
Concentration
ng/mL
Holding Time
(days)
0
2
3
7
14
21
30
IMPA
175
Average
%
Recovery
87.8
72.0
77.3
85.7
75.3
78.7
75.3
%
RSD
5
3
2
9
4
2
7
MPA
175
Average
%
Recovery
86.1
70.9
77.2
77.8
72.1
75.7
74.4
%
RSD
2
2
2
4
3
3
7
MPA-d3
175
Average
%
Recovery
78.9
72.7
73.8
75.6
71.6
72.3
73.5
%
RSD
3
1
4
3
3
4
7
EMPA
70
Average
%
Recovery
74.7
65.7
77.4
75.1
68.5
74.9
78.7
%
RSD
2
4
4
2
3
1
7
EHDMAP
175
Average
%
Recovery
106
24.4
20.8
28.2
29.5
33.4
48.8
%
RSD
3
2
3
15
3
3
54
PMPA
35
Average
%
Recovery
79.9
74.6
77.7
74.4
74.6
79.8
73.5
%
RSD
2
1
1
4
3
3
4
PMPA-13C6
175
Average
%
Recovery
79.5
71.3
75.4
74.4
74.8
70.6
72.6
%
RSD
2
1
1
1
2
2
4
RSD, relative standard deviation

-------
                                  Table 3. Method Parameters for Nerve Agent Degradation Products
LAMINATE
Analyte
IMPA
EMPA
EHDMAP
MPA
PMPA
DIMP
MDL*
ng/cm2t
0.042
0.050
0.067
0.065
0.017
-
ng/mL
4.2
5.0
6.7
6.5
1.7
-
MRL
ng/mL
25
10
25
25
5
-
*Final DL Study-8/12. ESI" ionization mode provided the method detection limit (MDL) and minimum reporting level (MRL) values.
See section 19.1 for complete DL data in both ionization modes.
tng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).

-------
 Table 4a. ESI (+) MRM Ion Transitions, Retention Time (RT) and Variable Mass Spectrometer
                                         Parameters
Analyte
DIMP
IMPA
EMPA
EHDMAP
MPA
DIMP-d14
MPA-d3
Cone
voltage
22
22
26
28
45
24
48
MRM mass transition
(parent — > product)
1 81 . 33 -» 139.25
139.29^96.80
125.22^96.82
154.29^125.82
97.25 -> 79.20
195.45^147.20
100.20^82.00
Collision
energy (eV)
7
18
12
16
15
7
16
RT*
(minutes)
7.6
6.6
4.0
3.2
1.8
7.6
1.8
Retention times should fall within 5% of the given value; otherwise re-analysis may be necessary.
 Table 4b. ESI (-) MRM Ion Transitions, Retention Time (RT) and Variable Mass Spectrometer
                                         Parameters
Analyte
IMPA
EMPA
EHDMAP
PMPA
MPA
PMPA-13C6
MPA-d3
Cone
voltage
30
26
30
38
45
34
37
MRM mass transition
(parent — > product)
137.18^95.00
123.10^94.95
152.17^78.92
179.20^95.00
95.06^78.95
185.22^94.99
98.00^78.80
Collision
energy (eV)
18
12
12
18
15
18
15
RT*
(minutes)
6.6
4.0
3.2
8.7
1.8
8.7
1.8
Retention times should fall within 5% of the given value; otherwise re-analysis may be necessary.

-------
                        Table 5. ESI (+) and (-) MS/MS Conditions
MS Parameter
Capillary Voltage
Cone Voltage
Extractor
RF Lens
Source Temperature
Desolvation Temperature
Desolvation Gas Flow
Cone Gas Flow
Low Mass Resolution 1
High Mass Resolution 1
Ion Energy 1
Entrance Energy
Collision Energy
Exit Energy
Low Mass Resolution 2
High Mass resolution 2
Ion Energy 2
Multiplier
Gas Cell Pirani Gauge
Inter-Channel Delay
Inter-Scan Delay
Repeats
Span
Dwell
Setting
4.3 kV
See Table 4a and b
2 Volts
0.2 Volts
150°C
350 °C
600 L/hr
50 L/hr
14.5
14.5
0.5
1
See Table 4a and b
1
15.0
15.0
0.5
-560
3.0x10"J Torr
0.005 seconds
0.005 seconds
1
0.1 Daltons
0.15 Seconds
                  Table 6. Liquid Chromatography Gradient Conditions*
Time
(min)
0
4
5
9
10
12
13
15
Flow
(jjL/min)
300
300
300
300
300
300
300
300
%
Solution Af
100
100
55
55
40
30
100
100
%
Solution Bft
0
0
45
45
60
70
0
0
 fA: Water (0.2% Formic Acid)
f fB: Acetonitrile (0.2% Formic Acid)
^Injection volume - 10 uL( recommended)
*Column Temperature: 30 ° C
*Autosampler Temperature: 15 °C
^Equilibration time: 2 minutes
*Column:, 100 mm x 2.1mm, 3|j,m particle size

-------
Table 7. Target Concentrations of Calibration
                                  Method
Standards Used During the Development of this
(ng/mL)
Analyte/Surrogate
DIMP
IMPA
EMPA
EHDMAP
PMPA
MPA
DIMP-d14
PMPA-13C6
MPA-d3
Level
1
5
25
10
25
5
25
5
25
25
Level
2
10
50
20
50
10
50
10
50
50
Level
3
20
100
40
100
20
100
20
100
100
Level
4
35
175
70
175
35
175
35
175
175
Level
5
50
250
100
250
50
250
50
250
250
Level
6
100
500
200
500
100
500
100
500
500
Level
7
150
750
300
750
150
750
150
750
750

-------
19  ATTACHMENTS
              19.1     Method Detection Limit Data and Calculations
              19.2    Preci sion and Accuracy
              19.3     Illustration depicting the wiping pattern on a 100 cm2 surface
                                               7

-------
     19.1
METHOD DETECTION LIMIT (MDL) DATA AND CALCULATIONS
     MDL Data for Seven Replicates for Nerve Agent Degradation Analytes
Analyte
IMPA
MPA
EM PA
EHDMAP
DIMP
PMPA
MPA-d3
DIMP-d14
PMPA-13C6
Analyte
IMPA
MPA
EM PA
EHDMAP
DIMP
PMPA
MPA-d3
DIMP-d14
PMPA-13C6
LAMINATE in ESI (+)
mode
Concentration 1*
Average
Recovery
(ng/mL)
54.6
65.3
19.5
39.5
ND
-
88.8
16.8
-
Average
Recovery
(ng/cm2)t
0.546
0.653
0.195
0.395
ND
-
0.888
0.168
-
%
Recovery
91.1
109
81.1
65.9
ND
-
88.8
84.1
-
%
Recovery
91.1
109
81.1
65.9
ND
-
88.8
84.1
-
%
RSD
8
6
8
8
-
-
6
4
-
%
RSD
8
6
8
8
-
-
6
4
-
LAMINATE in ESI (-)
mode
Concentration 1*
Average
Recovery
(ng/mL)
44.7
48.2
20.3
39.3
-
9.9
84.6
-
87.1
Average
Recovery
(ng/cm2)t
0.447
0.482
0.203
0.393
-
0.100
0.846
-
0.871
%
Recovery
74.5
80.3
84.5
65.4
-
82.9
84.6
-
87.1
%
Recovery
74.5
80.3
84.5
65.4
-
82.9
84.6
-
87.1
%
RSD
3
4
8
5
-
5
5
-
1
%
RSD
3
4
8
5
-
5
5
-
1
^Concentration 1 correlates to the following analyte concentrations: 60 ng/mL for IMPA, MPA, and
EHDMAP, 24 ng/mL for EMPA, and 12 ng/mL for DIMP and PMPA. Surrogate recovery concentrations
correspond to the following: 100 ng/mL for MPA-d3and PMPA-13C6, and 20 ng/mL for DIMP-d14.
tng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area
of the coupon (100 cm2).
                                            8

-------
     MDL Calculation for Seven Replicates for Nerve Agent Degradation Analytes
LAMINATE in ESI (+) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MDL
ng/cm2f
0.15
0.12
0.047
0.11
ND
ng/mL
15
12
4.7
11
ND
MRL
ng/mL
25
25
10
25
-
MDL, method detection limit; MRL, minimum reporting limit
tng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area
of the coupon (100 cm2).
LAMINATE in ESI (-) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MDL
ng/cm2f
0.042
0.065
0.049
0.067
0.017
ng/mL
4.2
6.5
4.9
6.7
1.7
MRL
ng/mL
25
25
10
25
5
MDL, method detection limit; MRL, minimum reporting limit
tng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area
of the coupon (100 cm2).

-------
     19.2     PRECISION AND ACCURACY

Concentration levels correspond to the following final concentrations on the surface: Concentration 1 is
the same as in Attachment 19.1 (60 ng/mL for IMPA, MPA, and EHDMAP, 24 ng/mL for EMPA, and 12
ng/mL for DIMP and PMPA. Surrogate recovery concentrations correspond to the following for
concentrations 1 and 2: 100 ng/mL for MPA-d3 and PMPA-13C6, and 20 ng/mL for DIMP-d14. Surrogate
recovery concentrations correspond to the following for concentrations 3 and 4: 300 ng/mL for MPA-d3
and PMPA-13C6, and 60 ng/mL for DIMP-di4).  Concentration 2 is calibration concentration level 3.
Concentrations 3 and 4 are 2.25 and 3 times Concentration 2.
    •   Table A. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Laminate Surfaces in ESI (+) Mode

    •   Table B. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Laminate Surfaces in ESI (-) Mode

    •   Table C. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Metal Surfaces in ESI (+) Mode

    •   Table D. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Metal Surfaces in ESI (-) Mode

    •   Table E. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Glass Surfaces in ESI (+) Mode

    •   Table F. Precision  and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Glass Surfaces in ESI (-) Mode

    •   Table G. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Painted Drywall Surfaces in ESI (+) Mode

    •   Table H. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Painted Drywall Surfaces in ESI (-) Mode

    •   Table I. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Vinyl Tile Surfaces in ESI (+) Mode

    •   Table J. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Vinyl Tile Surfaces in ESI (-) Mode

    •   Table K. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation
       Analytes on Treated Wood Surfaces in ESI (+) and ESI (-) Mode
                                            10

-------
P&A data for wipe analysis of nerve agent degradation analytes on surfaces.
Table A. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Laminate Surfaces in ESI (+)
LAMINATE in ESI (+) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Concentration 1
Average
Recovery
(ng/mL)*
54.6
65.3
19.5
39.5
ND
88.3
16.8
Average
Recovery
(ng/cm2)t
0.546
0.653
0.195
0.395
ND
0.883
0.168
%
Recovery
91.1
109
81.1
65.9
-
88.3
84.0
%
Recovery
91.1
109
81.1
65.9
-
88.8
84.0
%
RSD
8
6
8
8
-
3
4
%
RSD
8
6
8
8
-
3
4
Concentration 2
Average
Recovery
(ng/mL)*
97.6
115
40.6
72.6
ND
119
18.2
Average
Recovery
(ng/cm2)t
0.976
1.15
0.406
0.726
ND
1.19
0.182
%
Recovery
97.6
115
101
72.6
-
119
91.2
%
Recovery
97.6
115
101
72.6
-
119
91.2
%
RSD
10
8
7
9
-
15
13
%
RSD
10
8
7
9
-
15
13
Concentration 3
Average
Recovery
(ng/mL)*
219
277
90.9
127
ND
321
45.3
Average
Recovery
(ng/cm2)t
2.19
2.77
0.909
1.27
ND
3.21
0.453
%
Recovery
97.4
123
101
56.5
-
107
75.4
%
Recovery
97.4
123
101
56.5
-
107
75.4
%
RSD
6
5
4
6
-
8
8
%
RSD
6
5
4
6
-
8
8
Concentration 4
Average
Recovery
(ng/mL)*
283
263
106
210
ND
286
50.4
Average
Recovery
(ng/cm2)t
2.83
2.63
1.06
2.10
ND
2.86
0.504
%
Recovery
94.2
87.6
88.3
70.1
-
95.4
84.0
%
Recovery
94.2
87.6
88.3
70.1
-
95.4
84.0
%
RSD
18
5
3
9
-
2
8
%
RSD
18
5
3
9
-
2
8
                                                                                                                             Mode
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed  by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                 11

-------
Table B. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Laminate Surfaces in ESI (-) Mode
LAMINATE in ESI (-) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Concentration 1
Average
Recovery
(ng/mL)*
44.7
48.2
20.3
39.3
9.90
84.6
87.1
Average
Recovery
(ng/cm2)t
0.447
0.482
0.203
0.393
0.0990
0.846
0.871
%
Recovery
74.5
80.3
84.5
65.4
82.9
84.6
87.1
%
Recovery
74.5
80.3
84.5
65.4
82.9
84.6
87.1
%
RSD
3
4
8
5
5
5
1
%
RSD
3
4
8
5
5
5
1
Concentration 2
Average
Recovery
(ng/mL)*
69.2
58.3
38.2
71.4
18.1
68.7
94.6
Average
Recovery
(ng/cm2)t
0.692
0.583
0.382
0.714
0.181
0.687
0.946
%
Recovery
69.2
58.3
95.4
71.4
90.2
68.7
94.6
%
Recovery
69.2
58.3
95.4
71.4
90.2
68.7
94.6
%
RSD
9
6
7
10
3
7
14
%
RSD
9
6
7
10
3
7
14
Concentration 3
Average
Recovery
(ng/mL)*
146
132
86.5
128
41.5
170
245
Average
Recovery
(ng/cm2)t
1.46
1.32
0.865
1.28
0.415
1.70
2.45
%
Recovery
64.8
58.5
96.1
56.9
92.2
56.7
81.5
%
Recovery
64.8
58.5
96.1
56.9
92.2
56.7
81.5
%
RSD
5
5
4
7
5
3
4
%
RSD
5
5
4
7
5
3
4
Concentration 4
Average
Recovery
(ng/mL)*
234
243
108
217
52.3
261
276
Average
Recovery
(ng/cm2)t
2.34
2.43
1.08
2.17
0.523
2.61
2.76
%
Recovery
77.8
80.9
89.9
72.5
87.2
87.0
92.1
%
Recovery
77.8
80.9
89.9
72.5
87.2
87.0
92.1
%
RSD
1
3
2
5
3
2
2
%
RSD
1
3
2
5
3
2
2
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                12

-------
Table C. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Metal Surfaces in ESI (+) Mode
METAL in ESI (+) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Concentration 1
Average
Recovery
(ng/mL)*
65.1
50.0
19.6
16.8
ND
142
18.1
Average
Recovery
(ng/cm2)t
0.651
0.500
0.196
0.168
ND
1.42
0.181
%
Recovery
108
83.3
81.6
28.0
-
142
90.5
%
Recovery
108
83.3
81.6
28.0
-
142
90.5
%
RSD
20
11
12
29
-
4
6
%
RSD
20
11
12
29
-
4
6
Concentration 2
Average
Recovery
(ng/mL)*
162
86.8
32.5
71.5
ND
106
19.1
Average
Recovery
(ng/cm2)t
1.62
0.868
0.325
0.715
ND
1.06
0.191
%
Recovery
162
86.8
81.2
71.5
-
106
95.4
%
Recovery
162
86.8
81.2
71.5
-
106
95.4
%
RSD
17
15
6
14
-
6
5
%
RSD
17
15
6
14
-
6
5
Concentration 3
Average
Recovery
(ng/mL)*
353
182
74.8
138
ND
288
55.1
Average
Recovery
(ng/cm2)t
3.53
1.82
0.748
1.38
ND
2.88
0.551
%
Recovery
157
80.9
83.1
61.4
-
96.1
91.8
%
Recovery
157
80.9
83.1
61.4
-
96.1
91.8
%
RSD
16
17
4
30
-
3
4
%
RSD
16
17
4
30
-
3
4
Concentration 4
Average
Recovery
(ng/mL)*
236
213
85.7
75.9
ND
210
44.2
Average
Recovery
(ng/cm2)t
2.36
2.13
0.857
0.759
ND
2.10
0.442
%
Recovery
78.7
71.0
71.4
25.3
-
69.8
73.6
%
Recovery
78.7
71.0
71.4
25.3
-
69.8
73.6
%
RSD
9
13
7
55
-
5
5
%
RSD
9
13
7
55
-
5
5
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                13

-------
Table D. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Metal Surfaces in ESI (-) Mode
METAL in ESI (-) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Concentration 1
Average
Recovery
(ng/mL)*
42.7
28.2
20.0
13.0
8.80
89.0
66.9
Average
Recovery
(ng/cm2)t
0.427
0.282
0.200
0.130
0.0880
0.890
0.669
%
Recovery
71.2
47.0
83.4
21.7
73.0
89.0
66.9
%
Recovery
71.2
47.0
83.4
21.7
73.0
89.0
66.9
%
RSD
8
9
8
83
6
11
3
%
RSD
8
9
8
83
6
11
3
Concentration 2
Average
Recovery
(ng/mL)*
65.3
58.1
35.6
75.6
15.7
67.1
89.8
Average
Recovery
(ng/cm2)t
0.653
0.581
0.356
0.756
0.157
0.671
0.898
%
Recovery
65.3
58.1
89.0
75.6
78.6
67.1
89.8
%
Recovery
65.3
58.1
89.0
75.6
78.6
67.1
89.8
%
RSD
8
12
7
16
4
9
4
%
RSD
8
12
7
16
4
9
4
Concentration 3
Average
Recovery
(ng/mL)*
127
119
76.1
142
35.6
185
266
Average
Recovery
(ng/cm2)t
1.27
1.19
0.761
1.42
0.356
1.85
2.66
%
Recovery
56.2
52.8
84.6
62.9
79.2
61.5
88.6
%
Recovery
56.2
52.8
84.6
62.9
79.2
61.5
88.6
%
RSD
2
10
3
35
4
7
3
%
RSD
2
10
3
35
4
7
3
Concentration 4
Average
Recovery
(ng/mL)*
245
172
96.8
87.8
45.8
185
203
Average
Recovery
(ng/cm2)t
2.45
1.72
0.968
0.878
0.458
1.85
2.03
%
Recovery
81.8
57.3
80.7
29.3
76.3
61.8
67.6
%
Recovery
81.8
57.3
80.7
29.3
76.3
61.8
67.6
%
RSD
4
12
4
48
5
5
2
%
RSD
4
12
4
48
5
5
2
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                14

-------
Table E. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Glass Surfaces in ESI (+) Mode
GLASS in ESI (+) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Concentration 1
Average
Recovery
(ng/mL)*
141
71.4
25.0
44.1
ND
172
17.1
Average
Recovery
(ng/cm2)t
1.41
0.714
0.250
0.441
ND
1.72
0.171
%
Recovery
234
119
104
73.4
-
172
85.7
%
Recovery
234
119
104
73.4
-
172
85.7
%
RSD
13
11
5
6
-
6
10
%
RSD
13
11
5
6
-
6
10
Concentration 2
Average
Recovery
(ng/mL)*
210
105
33.4
121
ND
111
17.4
Average
Recovery
(ng/cm2)t
2.10
1.05
0.334
1.21
ND
1.11
0.174
%
Recovery
210
105
83.5
121
-
111
86.8
%
Recovery
210
105
83.5
121
-
111
86.8
%
RSD
10
7
7
4
-
10
11
%
RSD
10
7
7
4
-
10
11
Concentration 3
Average
Recovery
(ng/mL)*
292
253
84.0
217
ND
286
46.6
Average
Recovery
(ng/cm2)t
2.92
2.53
0.840
2.17
ND
2.86
0.466
%
Recovery
129
112
93.3
96.5
-
95.3
111
%
Recovery
129
112
93.3
96.5
-
95.3
111
%
RSD
31
4
6
5
-
5
8
%
RSD
31
4
6
5
-
5
8
Concentration 4
Average
Recovery
(ng/mL)*
444
347
113
270
ND
275
43.2
Average
Recovery
(ng/cm2)t
4.44
3.47
1.13
2.70
ND
2.75
0.432
%
Recovery
148
116
94.4
90.1
-
91.8
72.0
%
Recovery
148
116
94.4
90.1
-
91.8
72.0
%
RSD
8
4
2
3
-
2
7
%
RSD
8
4
2
3
-
2
7
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                15

-------
Table F. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Glass Surfaces in ESI (-) Mode
GLASS in ESI (-) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Concentration 1
Average
Recovery
(ng/mL)*
40.4
33.1
23.1
28.2
10.7
99.2
139
Average
Recovery
(ng/cm2)t
0.404
0.331
0.231
0.282
0.107
0.992
1.39
%
Recovery
67.4
55.2
96.0
47.1
88.8
99.2
139
%
Recovery
67.4
55.2
96.0
47.1
88.8
99.2
139
%
RSD
7
9
9
18
3
2
4
%
RSD
7
9
9
18
3
2
4
Concentration 2
Average
Recovery
(ng/mL)*
51.6
65.5
32.6
113
10.8
62.4
96.3
Average
Recovery
(ng/cm2)t
0.516
0.655
0.326
1.13
0.108
0.624
0.963
%
Recovery
51.6
65.5
81.4
113
54.0
62.4
96.3
%
Recovery
51.6
65.5
81.4
113
54.0
62.4
96.3
%
RSD
2
8
9
12
4
15
7
%
RSD
2
8
9
12
4
15
7
Concentration 3
Average
Recovery
(ng/mL)*
131
145
83.2
225
39.4
169
247
Average
Recovery
(ng/cm2)t
1.31
1.45
0.832
2.25
0.394
1.69
2.47
%
Recovery
58.4
64.3
92.4
100
87.5
56.2
82.3
%
Recovery
58.4
64.3
92.4
100
87.5
56.2
82.3
%
RSD
9
8
2
23
3
8
6
%
RSD
9
8
2
23
3
8
6
Concentration 4
Average
Recovery
(ng/mL)*
163
185
117
307
54.8
149
240
Average
Recovery
(ng/cm2)t
1.63
1.85
1.17
3.07
0.548
1.49
2.40
%
Recovery
54.3
61.5
97.3
102
91.4
49.7
79.9
%
Recovery
54.3
61.5
97.3
102
91.4
49.7
79.9
%
RSD
6
7
5
6
2
9
3
%
RSD
6
7
5
6
2
9
3
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                16

-------
Table G. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Painted Drywall Surfaces in ESI (+)
Mode
PAINTED DRYWALL in ESI (+) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Concentration 1
Average
Recovery
(ng/mL)*
40.3
59.0
18.9
33.9
ND
114
14.6
Average
Recovery
(ng/cm2)t
0.403
0.590
0.189
0.339
ND
1.14
0.146
%
Recovery
67.1
98.3
78.7
56.6
-
114
73.1
%
Recovery
67.1
98.3
78.7
56.6
-
114
73.1
%
RSD
6
28
14
10
-
7
13
%
RSD
6
28
14
10
-
7
13
Concentration 2
Average
Recovery
(ng/mL)*
67.0
110
32.5
50.0
ND
123
16.2
Average
Recovery
(ng/cm2)t
0.670
1.10
0.325
0.500
ND
1.23
0.162
%
Recovery
67.0
110.1
81.2
50.0
-
123
81.1
%
Recovery
67.0
110
81.2
50.0
-
123
81.1
%
RSD
3
11
7
6
-
8
9
%
RSD
3
11
7
6
-
8
9
Concentration 3
Average
Recovery
(ng/mL)*
184
223
64.1
163
ND
282
45.7
Average
Recovery
(ng/cm2)t
1.84
2.23
0.641
1.63
ND
2.82
0.457
%
Recovery
81.5
99.1
71.3
72.3
-
94.0
76.1
%
Recovery
81.5
99.1
71.3
72.3
-
94.0
76.1
%
RSD
6
5
3
3
-
7
8
%
RSD
6
5
3
3
-
7
8
Concentration 4
Average
Recovery
(ng/mL)*
236
285
78.6
231
ND
299
47.5
Average
Recovery
(ng/cm2)t
2.36
2.85
0.786
2.31
ND
2.99
0.475
%
Recovery
78.5
95.0
65.5
77.0
-
99.7
79.2
%
Recovery
78.5
95.0
65.5
77.0
-
99.7
79.2
%
RSD
3
8
5
2
-
8
9
%
RSD
3
8
5
2
-
8
9
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                17

-------
Table H. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Painted Drywall Surfaces in ESI (-)
Mode
PAINTED DRYWALL in ESI (-) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Concentration 1
Average
Recovery
(ng/mL)*
44.9
22.4
20.4
31.5
13.2
75.6
81.6
Average
Recovery
(ng/cm2)t
0.449
0.224
0.204
0.315
0.132
0.756
0.816
%
Recovery
74.8
37.4
85.2
52.6
110
75.6
81.6
%
Recovery
74.8
37.4
85.2
52.6
110
75.6
81.6
%
RSD
12
13
5
8
6
6
4
%
RSD
12
13
5
8
6
6
4
Concentration 2
Average
Recovery
(ng/mL)*
59.6
33.2
28.6
49.9
13.2
68.0
81.6
Average
Recovery
(ng/cm2)t
0.596
0.332
0.286
0.499
0.132
0.680
0.816
%
Recovery
59.6
33.2
71.5
49.9
65.8
68.0
81.6
%
Recovery
59.6
33.2
71.5
49.9
65.8
68.0
81.6
%
RSD
6
16
9
5
6
9
4
%
RSD
6
16
9
5
6
9
4
Concentration 3
Average
Recovery
(ng/mL)*
155
142
62.3
155
33.7
205
255
Average
Recovery
(ng/cm2)t
1.55
1.42
0.623
1.55
0.337
2.05
2.55
%
Recovery
68.8
63.1
69.3
68.8
74.9
68.3
85.0
%
Recovery
68.8
63.1
69.3
68.8
74.9
68.3
85.0
%
RSD
4
3
3
3
2
4
3
%
RSD
4
3
3
3
2
4
3
Concentration 4
Average
Recovery
(ng/mL)*
216
180
74.1
224
45.4
200
257
Average
Recovery
(ng/cm2)t
2.16
1.80
0.741
2.24
0.454
2.00
2.57
%
Recovery
72.1
30.1
61.7
74.7
75.7
66.6
85.8
%
Recovery
72.1
30.1
61.7
74.7
75.7
66.6
85.8
%
RSD
4
5
2
3
2
7
3
%
RSD
4
5
2
3
2
7
3
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                18

-------
Table I.  Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Vinyl Tile Surfaces in ESI (+) Mode
VINYL TILE in ESI (+) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Analyte
IMPA
MPA
EMPA
EHDMAP
DIMP
MPA-d3
DIMP-d14
Concentration 1
Average
Recovery
(ng/mL)*
57.3
87.5
22.2
42.2
ND
92.0
15.0
Average
Recovery
(ng/cm2)t
0.573
0.875
0.222
0.422
ND
0.920
0.150
%
Recovery
95.5
146
92.6
70.4
-
92.0
75.1
%
Recovery
95.5
146
92.6
70.4
-
92.0
75.1
%
RSD
16
21
6
7
-
9
8
%
RSD
16
21
6
7
-
9
8
Concentration 2
Average
Recovery
(ng/mL)*
95.0
108
32.8
53.6
ND
94.8
15.4
Average
Recovery
(ng/cm2)t
0.950
1.08
0.328
0.536
ND
0.948
0.154
%
Recovery
95.0
108
81.9
53.6
-
94.8
76.8
%
Recovery
95.0
108
81.9
53.6
-
94.8
76.8
%
RSD
26
8
13
7
-
8
7
%
RSD
26
8
13
7
-
8
7
Concentration 3
Average
Recovery
(ng/mL)*
191
136
62.9
164
ND
228
52.4
Average
Recovery
(ng/cm2)t
1.91
1.36
0.629
1.64
ND
2.28
0.524
%
Recovery
84.8
60.6
69.9
73.0
-
76.1
87.4
%
Recovery
84.8
60.6
69.9
73.0
-
76.1
87.4
%
RSD
9
13
3
3
-
2
4
%
RSD
9
13
3
3
-
2
4
Concentration 4
Average
Recovery
(ng/mL)*
290
208
111
231
ND
250
44.7
Average
Recovery
(ng/cm2)t
2.90
2.08
1.11
2.31
ND
2.50
0.447
%
Recovery
96.8
69.4
92.4
76.9
-
83.3
74.5
%
Recovery
96.8
69.4
92.4
76.9
-
83.3
74.5
%
RSD
11
6
5
5
-
2
10
%
RSD
11
6
5
5
-
2
10
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2; RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                 19

-------
Table J. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Vinyl Tile Surfaces in ESI (-) Mode
VINYL TILE in ESI (-) mode
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Analyte
IMPA
MPA
EMPA
EHDMAP
PMPA
MPA-d3
PMPA-13C6
Concentration 1
Average
Recovery
(ng/mL)*
47.0
41.1
19.0
47.9
8.2
86.9
81.1
Average
Recovery
(ng/cm2)t
0.470
0.411
0.190
0.479
0.0820
0.869
0.811
%
Recovery
78.4
68.5
79.2
79.8
68.5
86.9
81.1
%
Recovery
78.4
68.5
79.2
79.8
68.5
86.9
81.1
%
RSD
12
10
11
10
4
8
4
%
RSD
12
10
11
10
4
8
4
Concentration 2
Average
Recovery
(ng/mL)*
66.9
69.7
27.3
59.1
12.2
80.6
81.8
Average
Recovery
(ng/cm2)t
0.669
0.697
0.273
0.591
0.122
0.806
0.818
%
Recovery
66.9
69.7
68.2
59.1
60.8
80.6
81.8
%
Recovery
66.9
69.7
68.2
59.1
60.8
80.6
81.8
%
RSD
5
7
13
11
6
11
6
%
RSD
5
7
13
11
6
11
6
Concentration 3
Average
Recovery
(ng/mL)*
161
119
65.8
168
33.5
217
250.0
Average
Recovery
(ng/cm2)t
1.61
1.19
0.658
1.68
0.335
2.17
2.50
%
Recovery
71.8
52.9
73.2
74.7
74.4
72.3
83.3
%
Recovery
71.8
52.9
73.2
74.7
74.4
72.3
83.3
%
RSD
2
6
3
3
3
3
3
%
RSD
2
6
3
3
3
3
3
Concentration 4
Average
Recovery
(ng/mL)*
211
165
113
242
43.8
225
266
Average
Recovery
(ng/cm2)t
2.11
1.65
1.13
2.42
0.438
2.25
2.66
%
Recovery
70.4
55.1
94.3
80.6
73.0
74.9
88.7
%
Recovery
70.4
55.1
94.3
80.6
73.0
74.9
88.7
%
RSD
3
4
4
4
3
4
3
%
RSD
3
4
4
4
3
4
3
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2.
RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                20

-------
Table K. Precision and Accuracy Data for Wipe Analysis of Nerve Agent Degradation Analytes on Treated Wood Surfaces in ESI (+) and
ESI (-) Mode
Analyte
IMPA
MPA
EM PA
EHDMAP
DIMP
PMPA
MPA-d3
DIMP-d14
PMPA-13C6
Analyte
IMPA
MPA
EM PA
EHDMAP
DIMP
PMPA
MPA-d3
DIMP-d14
PMPA-13C6
TREATED WOOD in ESI (+) mode
Concentration 4
Average Recovery
(ng/ml_)*
20.9
5.70
ND
8.90
ND
-
257
55.5
-
Average Recovery
(ng/cm2)t
0.209
0.570
ND
0.0890
ND
-
2.57
0.555
-
% Recovery
7.00
1.90
ND
3.00
-
-
85.6
92.6
-
% Recovery
7.0
1.9
-
3.0
-
-
85.6
92.6
-
% RSD
17
121
-
14
-
-
5
7
-
% RSD
17
121
-
14
-
-
5
7
-
TREATED WOOD in ESI (-) mode
Concentration 4
Average Recovery
(ng/ml_)*
12.3
10.7
ND
8.1
ND
1.8
238
-
277
Average Recovery
(ng/cm2)t
0.123
0.107
ND
0.0810
ND
0.0180
2.38
-
2.77
% Recovery
4.10
3.60
ND
2.70
-
3.00
79.5
-
92.3
% Recovery
4.1
3.6
-
2.7
-
3.0
79.5
-
92.3
% RSD
9
16
-
21
-
16
8
-
3
% RSD
9
16
-
21
-
16
8
-
3
Concentration 1 is for low concentration of analyte on the surface, See attachments 19.1-19.2 for values; Concentration 2 is calibration
concentration level 3; Concentrations 3 and 4 are 2.25 and 3 times Concentration 2. RSD is relative standard deviation
* (n = 7 samples at each concentration)
t ng/cm2 calculation was performed by dividing the concentration spiked onto the surface by the test area of the coupon (100 cm2).
                                                                21

-------
      19.3     Illustration of wiping pattern on 100 cm surface
The analyte spike solution, containing the six analytes of interest, was added to the surface as shown in
19.3, allowed to completely dry (approximately 60-90 minutes depending on droplet size), and wiped
using wetted-cotton gauze wipes.
                                              22

-------
United States
Environmental Protection
Agency
PRESORTED STANDARD
 POSTAGE & FEES PAID
         EPA
   PERMIT NO. G-35
Office of Research and Development (8101R)
Washington, DC 20460

Official Business
Penalty for Private Use
$300

-------