THE ENVIRONMENTAL TECHNOLOGY VERIFICATION
                                        PROGRAM
   U.S. Environmental Protection Agency
                                                                       Batreiie
                                                                  Trtc Business o/ Innovation
                      ETV Joint Verification Statement
      TECHNOLOGY TYPE:  Rapid Toxicity Testing System

      APPLICATION:         Detecting Toxicity in Drinking Water
      TECHNOLOGY
      NAME:

      COMPANY:

      ADDRESS:
      WEB SITE:
      E-MAIL:
Toxi-Chromotest
Environmental Bio-Detection Products, Inc.
14 Abacus Road
Brampton, Ontario
CANADA L6T5B7

www.ebpi-kits.com
ebpi@ebpi-kits.com
PHONE: (905) 794-3274
FAX:    (905)794-2338
The U.S. Environmental Protection Agency (EPA) has established the Environmental Technology Verification
(ETV) Program to facilitate the deployment of innovative or improved environmental technologies through
performance verification and dissemination of information. The goal of the ETV Program is to further
environmental protection by accelerating the acceptance and use of improved and cost-effective technologies.
ETV seeks to achieve this goal by providing high-quality, peer-reviewed data on technology performance to
those involved in the design, distribution, financing, permitting, purchase, and use of environmental
technologies. Information and ETV documents are available at www.epa.gov/etv.

ETV works in partnership with recognized standards and testing organizations, with stakeholder groups
(consisting of buyers, vendor organizations, and permitters), and with individual technology developers. The
program evaluates the performance of innovative technologies by developing test plans that are responsive to
the needs of stakeholders, conducting field or laboratory tests (as appropriate), collecting and analyzing data,
and preparing peer-reviewed reports. All evaluations are conducted in accordance with rigorous quality
assurance (QA) protocols to ensure that data of known and adequate quality are generated and that the results
are defensible.

The Advanced Monitoring Systems (AMS) Center, one of six technology areas under ETV, is operated by
Battelle in cooperation with EPA's National Exposure Research  Laboratory. The AMS Center evaluated the
performance of the Environmental Bio-Detection Products, Inc. Toxi-Chromotest. This verification statement
provides a summary of the test results.

-------
VERIFICATION TEST DESCRIPTION

Rapid toxicity technologies use various biological organisms and chemical reactions to indicate the presence of
toxic contaminants. The toxic contaminants are indicated by a change or appearance of color or a change in
intensity. As part of this verification test, the Toxi-Chromotest was subjected to various concentrations of
contaminants such as industrial chemicals, pesticides, rodenticides, pharmaceuticals, nerve agents, and biological
toxins. Each contaminant was added to separate drinking water samples and analyzed. In addition to determining
whether the Toxi-Chromotest could detect the toxicity caused by each contaminant, its response to interfering
compounds, such as water treatment chemicals and by-products in clean drinking water, was evaluated.

The Toxi-Chromotest was evaluated by

 •  Endpoints and precision—color inhibition (as indicator of toxicity) with respect to that of the negative
    control for all concentration levels of contaminants and potential interfering compounds and consistency of
    the color change across replicate analyses

 •  Toxicity threshold for each contaminant—contaminant level at which higher concentrations generate
    inhibition significantly greater than the negative control and lower concentrations do not

 •  False positive responses—chlorination and chloramination by-product inhibition with respect to
    unspiked American Society for Testing and Materials Type II deionized water samples

 •  False negative responses—contaminants that, when present at lethal concentrations, did not produce any
    color inhibition with respect to the negative control

 •  Other performance factors (sample throughput, ease of use, reliability).

The Toxi-Chromotest was verified by analyzing a dechlorinated drinking water sample from Columbus, Ohio
(DDW), fortified with contaminants (at concentrations ranging from lethal levels to concentrations up to
1,000 times less than  the lethal dose) and interferences (metals possibly present as a result of the water
treatment processes).  Dechlorinated water was used because free chlorine kills the bacteria within the Toxi-
Chromotest reagent and can degrade the contaminants during storage. Inhibition results (endpoints) from four
replicates of each contaminant at each concentration level were evaluated to assess the ability of the Toxi-
Chromotest to detect toxicity, as well as to measure the precision of the Toxi-Chromotest results. The
response of the Toxi-Chromotest to possible interferents was evaluated by analyzing them at one-half of the
concentration limit recommended by the EPA's National Secondary Drinking Water Regulations guidance.
For analysis of by-products of the chlorination process, the unspiked DDW was analyzed because Columbus,
Ohio, uses chlorination as its  disinfectant procedure. For the analysis of by-products of the chloramination
process, a separate drinking water sample was obtained from the Metropolitan Water District of Southern
California (LaVerne,  California), which uses chloramination as its disinfection process. The samples were
analyzed after residual chlorine was removed using sodium thiosulfate. Sample throughput was measured
based on the number  of samples analyzed per hour. Ease of use and reliability were determined based on
documented observations of the operators.

Quality control samples included method blank samples, which consisted of American Society for Testing
and Materials Type II deionized water; positive control samples (fortified with mercuric chloride); and
negative control samples, which consisted of the unspiked DDW.

QA oversight of verification testing was provided by Battelle and EPA. Battelle  QA staff conducted a
technical systems audit, a performance evaluation audit, and a data quality audit of 10% of the test data.

This verification statement, the  full report on which it is based, and the test/QA plan for this verification test
are all available atwww.epa.gov/etv/centers/centerl.html.

-------
TECHNOLOGY DESCRIPTION

The following description of the Toxi-Chromotest is based on information provided by the vendor. This
technology description was not verified in this test.

The Toxi-Chromotest detects toxic substances in water, chemicals, pharmaceuticals, food, and body fluids.
The Toxi-Chromotest is a bacterial assay based on the ability of toxic materials and antibiotics to inhibit the
de novo synthesis of an inducible enzyme, B-galactosidase, in a strain of the bacteria, E. coll (K12 OR85). The
bacteria in the Toxi-Chromotest are exposed to stressing conditions and freeze dried. To test for toxicity, the
bacteria are mixed with a rehydration cocktail containing inducers of the enzyme B-galactosidase and factors
necessary for the recovery of the bacteria from their stressed condition. During the recovery phase, toxicants
present at sufficient concentrations penetrate the cell walls of the bacteria and inhibit the de novo synthesis of
the B-galactosidase. The rate of production of the induced enzyme is  detected by a reaction of the excreted
enzyme with a chromogenic substrate in the bacterial suspension that was exposed to the potential toxicant.
Toxic materials above threshold levels interfere with the production of the enzyme and decrease color
formation.

The Toxi-Chromotest kit includes a reaction mixture (the cocktail containing an inducer for the enzyme
B-galactosidase and co-factors required for the recovery of the bacteria from their stressed condition),
lyophilized bacteria, rehydration solution, a positive control (4 micrograms per milliliter of mercuric chloride
in water), a chromogenic substrate (blue chromogen cocktail, ready for use), and diluent for the positive
control and test samples. In addition, the Toxi-Chromotest kit contains three 96-well microtiter plates and
biohazard bags. The user must supply a micropipette for adding the test samples, rehydrated bacteria, and
chromogenic substrates to the test wells and an incubator in which the plates containing the bacteria are
allowed to recover and begin to produce the enzyme that reacts with the added chromogenic substrate. The
incubator must maintain a constant temperature of 37 °C during the 90 minute incubation period.

The Toxi-Chromotest is supplied in a 25- by 13- by 8- centimeter (cm) Styrofoam box that contains the 96-
well plates, the biohazard bags  for disposal of test materials, and all of the necessary reagents to carry out
three separate analytical test series. For field use, a 15-cm by 15-cm by 15-cm incubator can be supplied that
runs off a 12-volt battery or 120-volt alternating current.

The output from the Toxi-Chromotest can be measured in the laboratory by absorbance at 615 nanometers
using a plate reader. If the test is conducted in the field or a plate reader is not available (as during this test),
the results can be read by visually recording the intensity of blue color produced against an internally run set
of standards to obtain a relative toxicity reading. The standard Toxi-Chromotest kit, with reagent, bacteria,
and plates to run the tests in the three 96-well microtiter plates provided, sells for $375.

-------
VERIFICATION RESULTS
Parameter
Contaminants in
DDW
Potential
interferences in
DDW
False positive
response
False negative
response
Ease of use
Field portability
Throughput
Compound
Aldicarb
Botulinum toxin
complex B
Colchicine
Cyanide
Dicrotophos
Nicotine
Ricin
Soman
Thallium sulfate
VX
Interference
Aluminum
Copper
Iron
Manganese
Zinc
Lethal
Dose (LD)
Cone.
(mg/L)
260
0.3
240
250
1,400
2,800
15
1.4
2,800
2
Cone.
(mg/L)
0.5
0.6
0.15
0.25
2.5
Visual Observance of Color
Inhibition at Concentrations
Relative to the LD Concentration
LD
-
-
-
+
-
+
-
-
+
-
LD/10
-
-
-
+
-
-
-
-
+
-
LD/100
-
-
-
-
-
-
-
-
-
-
LD/1,000
-
-
-
-
-
-
-
-
+
-
Visual Observance of Color
Inhibition
-
-
-
-
-
Toxicity Threshold
(mg/L)
ND
ND
ND
25
ND
2,800
ND
ND
280
ND

The Toxi-Chromotest did not generate any false positive results to water containing chlorination
or chloramination by -products.
Aldicarb, botulinum toxin complex B, colchicine, dicrotophos, ricin, soman, and VX produced an
inhibition that was not visually distinguishable from the negative control at the lethal dose
concentrations.
The Toxi-Chromotest requires two 1.5-hour incubation periods. After bacteria rehydration, the
hydrated bacteria could be used only for one hour. In addition, the reaction of the Toxi-
Chromotest was observed visually, which was difficult when there were only slight variations in
color. No formal scientific training would be required to use the Toxi-Chromotest.
Overall the Toxi-Chromotest was easy to transport to the field and, with an incubator
warmed ahead of time, was deployed in a matter of minutes. Results were obtained within 3
to 4 hours of starting the test. Each Toxi-Chromotest kit contained materials to process three
96-well plates.
The number of samples that can be processed depends on the number of replicates per sample
and the number of dilutions per sample that are processed on each 96-well plate. One plate can be
taken through the procedure in 3 to 4 hours. Each Toxi-Chromotest kit contained materials to
process samples filling three 96-well plates.
 + = Visually distinguishable color inhibition from that of the negative control was observed.
 - = Visually distinguishable color inhibition from that of the negative control was not observed

-------
Original signed by Gregory A. Mack
Gregory A. Mack                          Date
Vice President
Energy, Transportation, and Environment Division
Battelle
6/22/06   Original signed by Andrew P. Avel
          Andrew P. Avel
          Acting Director
          National Homeland Security Research Center
          Office of Research and Development
          U.S. Environmental Protection Agency
8/7/06
Date
     NOTICE: ETV verifications are based on an evaluation of technology performance under specific,
     predetermined criteria and the appropriate quality assurance procedures. EPA and Battelle make no expressed or
     implied warranties as to the performance of the technology and do not certify that a technology will always
     operate as verified. The end user is solely responsible for complying with any and all applicable federal, state,
     and local requirements. Mention of commercial product names does not imply endorsement.

-------