NATIONALFUNCTIONALGUIDELINES
                     for Inorganic Superfund Data Review
&EPA
Office of Superfund Remediation and Technology Innovation (OSRTI)
United States Environmental Protection Agency (EPA)
Washington, DC 20460
OSWER 9355.0-131
EPA-540-R-013-001
AUGUST 2014

-------
                                This page is intentionally left blank.
August 2014

-------
                                           NOTICE

The policies and procedures set forth here are intended as guidance to the United States Environmental
Protection Agency (EPA) and other governmental employees. They do not constitute rule making by the
EPA, and may not be relied upon to create a substantive or procedural right enforceable by any other
person.  The Government may take action that is at variance with the policies and procedures in this
manual.
This document can be obtained from the EPA's Contract Laboratory Program (CLP) website at:
                     http://www.epa.gov/superfund/programs/clp/guidance.htm
August 2014                                                                                 iii

-------
                                This page is intentionally left blank.
August 2014                                                                                      iv

-------
                                 TABLE OF CONTENTS
LIST OF TABLES	vii
ACRONYMS AND ABBREVIATIONS	ix
I.    Terminology	ix
II.   Target Analyte List	xi

INTRODUCTION	1
I.    Purpose of Document	1
II.   Limitations of Use	1
III.   Document Organization	1
IV.   For Additional Information	1

PART A: GENERAL DATA REVIEW	3
I.    Preliminary Review	5
II.   Data Qualifier Definitions	6
III.   Data Review Narrative	7
PART B: METHOD-SPECIFIC DATA REVIEW	9
ICP-AES DATA REVIEW	11
Example Analytical Sequence	13
I.    Preservation and Holding Times	15
II.   Calibration	17
III.   Blanks	20
IV.   Interference Check Sample	24
V.   Laboratory Control Sample	27
VI.   Duplicate Sample Analysis	30
VII.  Spike Sample Analysis	33
VIII. Serial Dilution	36
IX.   Regional Quality Assurance and Quality Control	38
X.   Overall Assessment of Data	39
XI.   Calculations	41

ICP-MS DATA REVIEW	43
Example Analytical Sequence	45
I.    Preservation and Holding Times	47
II.   Tune Analysis	49
III.   Calibration	51
IV.   Blanks	54
V.   Interference Check Sample	58
VI.   Laboratory Control Sample	61
VII.  Duplicate Sample Analysis	63
VIII. Spike Sample Analysis	66
IX.   Serial Dilution	69
X.   Internal Standards	71
XI.   Regional Quality Assurance and Quality Control	73
XII.  Overall Assessment of Data	74
XIII. Calculations	76
August 2014

-------
MERCURY DATA REVIEW	77
Example Analytical Sequence	79
I.    Preservation and Holding Times	81
II.   Calibration	83
III.   Blanks	86
IV.   Duplicate Sample Analysis	90
V.   Spike Sample Analysis	93
VI.   Regional Quality Assurance and Quality Control	95
VII.  Overall Assessment of Data	96
VIII. Calculations	98

CYANIDE DATA REVIEW	99
Example Analytical Sequence	101
I.    Preservation and Holding Times	103
II.   Calibration	105
III.   Blanks	109
IV.   Duplicate Sample Analysis	113
V.   Spike Sample Analysis	116
VI.   Regional Quality Assurance and Quality Control	119
VII.  Overall Assessment of Data	120
VIII. Calculations	122
APPENDIX A: GLOSSARY	A-l
APPENDIX B: INORGANIC DATA REVIEW SUMMARY	B-l
August 2014                                                                              vi

-------
Inorganic Data Review
                                     LIST OF TABLES

Table 1.   Data Qualifiers and Definitions	6
Table 2.   Preservation and Holding Time Actions for ICP-AES Analysis	16
Table 3.   Acceptance Criteria for ICV and CCV Standards for ICP-AES Analysis	17
Table 4.   Calibration Actions for ICP-AES Analysis	19
Table 5.   Blank Actions for ICP-AES Analysis	22
Table 6.   Interference Check Actions for ICP-AES Analysis	26
Table 7.   LCS Actions for ICP-AES Analysis	29
Table 8.   Duplicate Sample Actions for ICP-AES Analysis	32
Table 9.   Spike Sample Actions for ICP-AES Analysis	35
Table 10.  Serial Dilution Actions for ICP-AES Analysis	37
Table 11.  Preservation and Holding Time Actions for ICP-MS  Analysis	48
Table 12.  ICP-MS Tune Actions for ICP-MS Analysis	50
Table 13.  Acceptance Criteria for ICV and CCV Standards for ICP-MS Analysis	51
Table 14.  Calibration Actions for ICP-MS Analysis	53
Table 15.  Blank Actions for ICP-MS Analysis	56
Table 16.  Interference Check Actions for ICP-MS Analysis	60
Table 17.  LCS Actions for ICP-MS Analysis	62
Table 18.  Duplicate Sample Actions for ICP-MS Analysis	65
Table 19.  Spike Sample Actions for ICP-MS Analysis	68
Table 20.  Serial Dilution Actions for ICP-MS Analysis	70
Table 21.  Internal Standard Actions for ICP-MS Analysis	72
Table 22.  Preservation and Holding Time Actions for Mercury Analysis	82
Table 23.  Acceptance Criteria for ICV and CCV Standards for Mercury Analysis	83
Table 24.  Calibration Actions for Mercury Analysis	85
Table 25.  Blank Actions for Mercury Analysis	89
Table 26.  Duplicate Sample Actions for Mercury Analysis	92
Table 27.  Spike Sample Actions for Mercury Analysis	94
Table 28.  Preservation and Holding Time Actions for Cyanide  Analysis	104
Table 29.  Acceptance Criteria for ICV and CCV Standards for Cyanide Analysis	105
Table 30.  Calibration Actions for Cyanide Analysis	108
Table 31.  Blank Actions for Cyanide Analysis	Ill
Table 32.  Duplicate Sample Actions for Cyanide Analysis	115
Table 33.  Spike Sample Actions for Cyanide Analysis	118
August 2014
VII

-------
Inorganic Data Review
                                 This page is intentionally left blank.
August 2014                                                                                     viii

-------
Inorganic Data Review
I.   Terminology
    The following
    definition, see
    CCB
    CCS
    CCV
    CLP
    COR
    CRQL
    CSF
    DF
    DQO
    EDM
    EPA
    EXES
    ICB
    ICP
    ICP-AES
    ICP-MS
    ICS
    ICV
    LCS
    LEB
    MDL
    NFG
    %D
    %R
    %RSD
    %Solids
    PE
    QA
    QAPP
    QC
    RPD
           ACRONYMS AND ABBREVIATIONS

acronyms and abbreviations may be found throughout this document. For further
Appendix A: Glossary at the end of the document.
Continuing Calibration Blank
Contract Compliance Screening
Continuing Calibration Verification
Contract Laboratory Program
Contracting Officer Representative
Contract Required Quantitation Limit
Complete SDG File
Dilution Factor
Data Quality Objective
EXES Data Manager
United States Environmental Protection Agency
Electronic Data Exchange and Evaluation System
Initial Calibration Blank
Inductively Coupled Plasma
Inductively Coupled Plasma - Atomic Emission Spectroscopy
Inductively Coupled Plasma - Mass Spectrometry
Interference Check Sample
Initial Calibration Verification
Laboratory Control Sample
Leachate Extraction Blank
Method Detection Limit
National Functional Guidelines
Percent Difference
Percent Recovery
Percent Relative Intensity
Percent Relative Standard Deviation
Percent Solids
Performance Evaluation
Quality Assurance
Quality Assurance Project Plan
Quality Control
Relative Percent Difference
August 2014
                                                                          IX

-------
Inorganic Data Review
    SDG        Sample Delivery Group
    SEDD       Staged Electronic Data Deliverable
    SMO        Sample Management Office
    SOP        Standard Operating Procedure
    SOW        Statement of Work
    SPLP        Synthetic Precipitation Leaching Procedure
    TAL        Target Analyte List
    TCLP       Toxicity  Characteristic Leaching Procedure
    TDS        Total Dissolved Solids
    TOC        Total Organic Carbon
    TR/COC    Traffic Report/Chain of Custody
    TSS        Total Suspended Solids
August 2014

-------
Inorganic Data Review
II.  Target Analyte List
    The EPA Contract Laboratory Program (CLP) Statement of Work (SOW) for Inorganic Superfund
    Methods (Multi-Media, Multi-Concentration) ISM02.2 applies CLP analytical methods for the
    isolation, detection, and quantitation of the following target analytes and parameter:
   Al
   Sb
   As
   Ba
   Be
   Cd
   Ca
   Cr
   Co
   Cu
   CN
   Fe
   Pb
   Mg
   Mn
   Hg
   Ni
   K
   Se
   Ag
   Na
   Tl
   V
   Zn
   Hardness
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Cyanide
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
August 2014
                                                                          XI

-------
Inorganic Data Review
                                This page is intentionally left blank.
August 2014                                                                                     xii

-------
Inorganic Data Review	Introduction

                                       INTRODUCTION

I.   Purpose of Document
    This document contains guidance to aid the data reviewer in determining the usability of analytical
    data generated using the United States Environmental Protection Agency (EPA) Contract Laboratory
    Program (CLP) Statement of Work (SOW) for Inorganic Superfund Methods (Multi-Media, Multi-
    Concentration) ISM02.2. The SOW includes metals by Inductively Coupled Plasma-Atomic
    Emission Spectroscopy (ICP-AES), metals by Inductively Coupled Plasma-Mass Spectrometry (ICP-
    MS), mercury, and cyanide analytical methods.
    The guidelines presented in this document are designed to assist the reviewer in evaluating (a)
    whether the analytical data meet the technical and Quality Control (QC) criteria specified in the
    SOW, and (b) the usability and extent of bias of any data that do not meet these criteria. This
    document contains definitive guidance in areas, such as blanks, calibration verification standards,
    Interference Check Samples (ICSs), QC audit samples, and instrument performance checks (e.g.,
    tuning), in which performance is fully under a laboratory's control.  General guidance is provided to
    aid the reviewer in making subjective judgments regarding the use of data that are affected by site
    conditions (e.g., sample matrix effects) and do not meet SOW-specified requirements.

II.  Limitations of Use
    This guidance is specific to the review of analytical data generated using CLP SOW ISM02.2.  It
    applies to the current version of the SOW, as well as future versions that contain editorial changes.
    To use this document effectively, the reviewer should have an understanding of the analytical
    methods and a general overview of the Sample Delivery Group (SDG) or Case  at hand.  This
    guidance is not appropriate  for use in conducting contract compliance reviews and should be used
    with caution in reviewing data generated using methods other than CLP SOW ISM02.2, although the
    general types of QC checks, the evaluation procedures, and the  decisions made  after consideration of
    the evaluation criteria may be applicable to data from any similar method.
    While this document is a valuable aid in the data review process, other sources  of guidance and
    information, along with professional judgment, are useful in determining the ultimate usability of the
    data. This is particularly critical in those cases where data do not meet  SOW-specified technical and
    QC criteria. To make the appropriate judgments, the reviewer needs to gain a complete
    understanding of the intended use of the data and is strongly encouraged to establish a dialogue with
    the data user prior to and following data review, to discuss usability issues and  resolve questions
    regarding the review.

III. Document Organization
    Following this introduction, the document is presented in two major parts: Part  A - General Data
    Review, which applies to all methods; and Part B - Method-Specific Data Review. In Part B, each
    method is addressed individually in a stand-alone format.  A complete list of acronyms used in this
    document appears preceding this introduction, and a Glossary is appended as Appendix A.
IV. For Additional Information
    For additional information regarding the CLP and the services it provides, refer to the CLP website at
    http://www.epa.gov/superfund/programs/clp/index.htm
August 2014

-------
Inorganic Data Review
                                This page is intentionally left blank.
August 2014

-------
Inorganic Data Review                                                         General
                        PART A: GENERAL DATA REVIEW
August 2014

-------
Inorganic Data Review                                                                      General
                                This page is intentionally left blank.
August 2014

-------
Inorganic Data Review                                                                    General

I.   Preliminary Review
    A preliminary review should be performed on the data, prior to embarking on the method-specific
    review (see Part B).  During this process, the reviewer should compile the necessary data package
    elements to ensure that all of the information needed to determine data usability is available. The
    preliminary review also allows the reviewer to obtain an overview of the Case or Sample Delivery
    Group (SDG) under review.
    The initial review should include, but is not limited to, verification of the exact number of samples,
    their assigned number and matrices, and the Contractor laboratory name.  It should take into
    consideration all the  documentation specific to the sample data package, which may include Modified
    Analysis requests, Traffic Report/Chain of Custody (TR/COC) Record, SDG Narrative, and other
    applicable documents.
    The reviewer should be aware that minor modifications to the  Statement of Work (SOW) that have
    been made through a Modified Analysis request, to meet site-specific requirements, could affect
    certain validation criteria such as Contract Required Quantitation Limits (CRQLs) and Target Analyte
    Lists (TALs). Therefore, these modifications should be applied during the method-specific review
    (Part B) process.
    The Cases or SDGs routinely have unique field quality control (QC) samples that may affect the
    outcome of the review. These include field blanks, field duplicates, and Performance  Evaluation (PE)
    samples which must be identified in the sampling records. The reviewer should verify that the
    following items are identified in the sampling records (e.g., TR/COC Records, field logs, and/or
    contractor tables):
    1.  The United States Environmental Protection Agency (EPA) Region where the samples were
       collected; and
    2.  The complete list of samples with information on:
       a.   Sample matrix
       b.   Field blanks (if applicable)
       c.   Field duplicates (if applicable)
       d.   Field spikes  (if applicable)
       e.   PE samples (if applicable)
       f   Sampling dates
       g.   Sampling times
       h.   Shipping dates

       i.   Preservatives
       j.   Types of analysis
       k.   Contractor laboratory

    The laboratory's SDG Narrative is another source of general information which includes notable
    problems with matrices; insufficient sample volume for analysis  or reanalysis; samples received in
    broken containers; preservation information; and unusual events. The reviewer should also inspect
    any email or telephone/communication logs in the data package detailing any discussion of sample
    and/or analysis issues between the laboratory, the Contract Laboratory Program (CLP) Sample
    Management Office (SMO), and the EPA Region.
    The reviewer should also have a copy of the Quality Assurance Project Plan (QAPP),  or similar
    document, for the project for which the  samples were analyzed, to assist in the determination of final
    usability of the analytical data. The reviewer should contact the  appropriate Regional  Laboratory

August 2014                                                                                   5

-------
Inorganic Data Review
General
    Contracting Officer Representative (COR) to obtain copies of the QAPP and relevant site
    information.

    For data obtained through the CLP, the Staged Electronic Data Deliverable (SEDD) generated by the
    CLP laboratories is subjected to the following reviews via the Electronic Data Exchange and
    Evaluation System (EXES): 1) automated data assessment for Contract Compliance Screening (CCS)
    based on the technical and QC criteria in CLP SOW ISM02.2, and 2) automated data validation based
    on the criteria in the EPA CLP National Functional Guidelines (NFG) for Inorganic Superfund Data
    Review.  In addition, completeness checks are manually performed on the hardcopy data. The
    automated CCS results and hardcopy data issues are subsequently included in a CCS defect report
    that is provided to the laboratory. The laboratory may then submit a reconciliation package for any
    missing items or to correct non-compliant data identified in the report.  The automated data validation
    results are summarized in criteria-based NFG reports that are provided to the EPA Regions. The data
    reviewer can access the CCS and NFG reports through the EXES Data Manager (EDM) via the SMO
    Portal and may use them in determining data usability.
    For more information about EXES and EDM, refer to the following CLP website:
    http://www.epa.gov/superfund/programs/clp/data assessment.htm

    For access to the SMO Portal, refer to the following CLP website to contact the Regional Laboratory
    COR from the Region where the data review is being performed and to obtain the necessary username
    and password information:

    http: //www. epa.gov/superfund/programs/clp/contacts .htm
    For concerns or questions regarding the data package, contact the Regional Laboratory COR from the
    Region where the samples were collected.

II.  Data Qualifier Definitions
    The following definitions provide brief explanations of the national qualifiers assigned to results
    during the data review process. The reviewer should use these qualifiers as applicable. If the
    reviewer chooses to use additional qualifiers, a complete explanation of those qualifiers shall
    accompany the data review.

                            Table 1. Data Qualifiers and Definitions
Data
Qualifier
U
J
J+
J-
UJ
R
Definition
The analyte was analyzed for, but was not detected above the level of the reported sample
quantitation limit.
The result is an estimated quantity. The associated numerical value is the approximate
concentration of the analyte in the sample.
The result is an estimated quantity, but the result may be biased high.
The result is an estimated quantity, but the result may be biased low.
The analyte was analyzed for, but was not detected. The reported quantitation limit is
approximate and may be inaccurate or imprecise.
The data are unusable. The sample results are rejected due to serious deficiencies in
meeting QC criteria. The analyte may or may not be present in the sample.
August 2014

-------
Inorganic Data Review                                                                    General

III. Data Review Narrative
    The reviewer should complete a Data Review Narrative that includes comments that address the
    problems identified during the review process and state the limitations of the data associated with a
    Case or SDG. The CLP Sample Numbers, analytical methods, extent of the problem(s), and assigned
    qualifiers should also be listed in the document.
    The Data Review Narrative, including the Inorganic Data Review Summary form (see Appendix B),
    must be provided together with the laboratory data to the appropriate recipient(s). A copy of the Data
    Review Narrative should also be submitted to the Regional Laboratory COR assigned oversight
    responsibility for the Contractor laboratory.
August 2014

-------
Inorganic Data Review
                                This page is intentionally left blank.
August 2014

-------
Inorganic Data Review
                     PART B: METHOD-SPECIFIC DATA REVIEW
August 2014

-------
Inorganic Data Review
                                This page is intentionally left blank.
August 2014                                                                                    10

-------
Inorganic Data Review                                                                 ICP-AES
                                  ICP-AES DATA REVIEW
The inorganic data requirements for Inductively Coupled Plasma-Atomic Emission Spectroscopy (ICP-
AES) to be reviewed during validation are listed below:
Example Analytical Sequence	13
I.    Preservation and Holding Times	15
II.   Calibration	17
III.   Blanks	20
IV.   Interference Check Sample	24
V.   Laboratory Control Sample	27
VI.   Duplicate Sample Analysis	30
VII.  Spike Sample Analysis	33
VIII. Serial Dilution	36
IX.   Regional Quality Assurance and Quality Control	38
X.   Overall  Assessment of Data	39
XI.   Calculations	41
August 2014                                                                                11

-------
Inorganic Data Review                                                                  ICP-AES
                               This page is intentionally left blank.
August 2014                                                                                  12

-------
Inorganic Data Review                                                           ICP-AES

                              Example Analytical Sequence
This is an example of an analytical sequence:

S##
S##
s##
s##
s##
s##
ICV
ICB
ICSA
ICSAB
ccv###
CCB###
samples
ccv###
CCB###
samples
ccv###
CCB###, etc.
* Suffix ## and ### are as specified in Exhibit B of the Statement of Work (SOW).
August 2014                                                                         13

-------
Inorganic Data Review                                                                  ICP-AES
                               This page is intentionally left blank.
August 2014                                                                                  14

-------
Inorganic Data Review                                                                  ICP-AES

                            I.       Preservation and Holding Times
A.  Review Items
    Form 1-IN, Form 12-IN, Traffic Report/Chain of Custody (TR/COC) documentation, Form DC-1,
    raw data, and the Sample Delivery Group (SDG) Narrative checking for: pH; shipping container
    temperature; holding time; and other sample conditions.

B.  Objective
    The objective is to determine the validity of the analytical results based on the sample conditions and
    the holding time of the sample.

C.  Criteria
    1.   The technical holding time is determined from the date of collection, or the date Toxicity
        Characteristic Leaching Procedure (TCLP) or Synthetic Precipitation Leaching Procedure (SPLP)
        extraction is complete, to the date of analysis.
    2.   The technical holding time criteria for aqueous/water samples and leachate samples from TCLP
        or SPLP is 180 days, preserved (with nitric acid) to pH < 2.
    3.   The technical holding time criteria for soil/sediment samples is 180 days, based on the technical
        holding time criteria for aqueous/water samples.
    4.   The technical holding time criteria for wipe samples is 180 days, based on the technical holding
        time criteria for aqueous/water samples.
    5.   Soil/sediment samples shall be maintained at < 6°C (but not frozen) from the time of collection
        until receipt at the laboratory. All aqueous/water and soil/sediment samples must be stored at
        < 6°C (but not frozen) from the time of sample receipt until digestion. The TCLP and SPLP
        leachates must be stored at < 6°C (but not frozen) from the time of the leaching procedure
        completion until digestion. Wipe samples should be  maintained at room temperature until
        preparation.

D.  Evaluation
    Establish technical holding times by comparing the sampling date(s) on the TR/COC documentation
    with the  dates of analysis on Form 12-IN and the raw data; also consider using information in the
    Complete SDG File (CSF), as it may be helpful in the assessment. Verify that the analysis dates on
    the Form 12-IN and the raw data are identical. Review the SDG Narrative and raw data preparation
    logs to determine if samples were properly preserved. If there is an indication of problems with the
    samples, the sample integrity may be compromised. Use professional judgment to evaluate the effect
    of the problem on the sample results.
E.  Action
    NOTE:     Apply the action to each field  sample for which the preservation or holding time criteria
               was not met.
    1.   If the pH of aqueous/water samples is  > 2 at the time of sample receipt, determine if the
        laboratory adjusted the pH of the sample to < 2 at the time of sample receipt. Also determine if
        the laboratory adjusted the pH to < 2 for the TCLP and SPLP leachates after completion of the
        leaching procedure.  If not, use  professional judgment to qualify the samples based on the pH of
        the sample and the chemistry of the metal(s) of interest. Detects should be qualified as estimated
        low (J-) and non-detects as unusable (R).
    2.   If soil/sediment samples are not maintained at < 6°C  (but not frozen) from the time of collection
        until receipt at the laboratory, detects should be qualified as estimated low (J-) and non-detects as
        unusable (R).
August 2014                                                                                  15

-------
Inorganic Data Review
ICP-AES
    3.  If technical holding times are exceeded, use professional judgment to determine the reliability of
       the data, based on the magnitude of the additional time compared to the technical requirement and
       whether the samples were properly preserved.  The expected bias would be low. Detects should
       be qualified as estimated low (J-) and non-detects as unusable (R).
    4.  Due to limited information concerning holding times for soil/sediment and wipe samples, use
       discretion when deciding whether to  apply the aqueous/water holding time criteria to
       soil/sediment and wipe samples. If they are applied, document this action in the Data Review
       Narrative.
    5.  If aqueous/water and soil/sediment samples are received with shipping container temperatures
       > 10°C, use professional judgment to determine the reliability of the data, or qualify detects as
       estimated (J) and non-detects as estimated (UJ).
    6.  When the holding times are exceeded, annotate any possible consequences for the analytical
       results in the Data Review Narrative, and note it for Regional Laboratory Contracting Officer
       Representative (COR) action.

             Table 2. Preservation and Holding Time Actions for ICP-AES Analysis
Criteria
Aqueous/water samples received with pH > 2 and pH not
adjusted
TCLP/SPLP leachate samples with pH >2 and pH not adjusted
Soil/sediment samples not maintained at < 6°C (but not frozen)
from time of collection until receipt at the laboratory
Technical Holding Time:
Aqueous/water and TCLP/SPLP leachate samples > 180 days
Technical Holding Time:
Soil/sediment and wipe samples > 180 days
Samples received > 10°C*
Action
Detect
Use professional
judgment
J-
Use professional
judgment
J-
Use professional
judgment
J-
J-
J-
Use professional
judgment
J
Non-detect
Use professional
judgment
R
Use professional
judgment
R
Use professional
judgment
R
R
R
Use professional
judgment
UJ
    *  For samples received with shipping container temperatures > 10°C, Regional policy or project
       Data Quality Objectives (DQO) may allow the use of higher temperature criteria before assessing
       any actions for the affected samples.
August 2014
      16

-------
Inorganic Data Review                                                                    ICP-AES

                                       II.     Calibration
A.  Review Items
    Form 2-IN, Form 12-IN, Form 15-IN, Form 16-IN, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on initial calibration and
    calibration verification.

C.  Criteria
    1.  Initial Calibration
       The instruments shall be successfully calibrated each time the instrument is set up and after
       Continuing Calibration Verification (CCV) failure.  The calibration date and time shall be
       included in the raw data.
       a.  A blank and at least five calibration standards shall be used to establish each calibration
           curve. At least one of these standards shall be at or below the Contract Required Quantitation
           Limit (CRQL) but  above the Method  Detection Limit (MDL). All measurements shall be
           within the instrument working range where the interelement correction factors are valid. A
           minimum of three replicate exposures are required for standardization, for all Quality Control
           (QC), and sample analyses. The average result of all the multiple exposures for the
           standardization, QC, and sample analyses shall be used. The calibration curve shall be fitted
           using linear regression or weighted linear regression. The curve may be forced through zero.
           The curve must have a correlation coefficient > 0.995.  The calculated  percent differences
           (%Ds) for all of the non-zero standards must be within ±30% of the true value of the
           standard. The y-intercept of the curve must be less than the CRQL.
    2.  Initial and Continuing Calibration Verification
       The acceptance criteria for the Initial Calibration Verification (ICV) and CCV standards are
       presented in Table 3:

         Table 3. Acceptance Criteria for ICV and CCV Standards for ICP-AES Analysis
Analytical Method
ICP-AES
Inorganic Analytes
Metals
ICV/CCV Low Limit
(% of True Value)
90
ICV/CCV High Limit
(% of True Value)
110
       a.  Initial Calibration Verification

           1)  Immediately after each system has been calibrated, the accuracy of the initial calibration
               must be verified and documented for each target analyte by the analysis of an ICV
               solution(s). If the ICV Percent Recovery (%R) falls outside of the control limits, the
               analysis should be terminated, the problem corrected, the instrument recalibrated, and all
               affected samples reanalyzed.
           2)  Only if the ICV is not available from the United States Environmental Protection Agency
               (EPA), analyses shall be conducted using a certified solution of the analytes from an
               independent commercial standard source, at a concentration level other than that used for
               instrument calibration, but within the calibrated range.
           3)  The ICV solution shall be analyzed at each analytical wavelength used for analysis.
       b.  Continuing Calibration Verification

           1)  To ensure  accuracy during the course of each analytical sequence, the CCV shall be
               analyzed and reported for each wavelength used for the analysis of each analyte.

August 2014                                                                                    17

-------
Inorganic Data Review                                                                   ICP-AES

           2)  The CCV standard shall be analyzed at a frequency of every two hours during an
               analytical sequence.  The CCV standard shall also be analyzed at the beginning of the
               analytical sequence, and again after the last analytical sample.
           3)  The analyte concentration(s) in the CCV standard(s) shall be different than the
               concentration used for the ICV, and at a concentration equivalent to the mid-level of their
               respective calibration curves.
           4)  The same CCV standard solution shall be used throughout the analysis for an SDG.
           5)  The CCV shall be analyzed in the same fashion as an actual sample.  If the %R of the
               CCV was outside of the control limits, the analysis should be terminated, the problem
               corrected, the instrument recalibrated, and all analytical samples analyzed since the last
               compliant CCV reanalyzed.
D.  Evaluation
    1.  Verify that the instrument was calibrated each time the instrument was set up, utilizing a blank
       and at least five calibration standards, one of which was at or below the CRQL but above the
       MDL.

    2.  Confirm that the measurements were within the working calibration range, and were the average
       result of at least three replicate exposures.
    3.  Verify that the ICV and CCV standards were analyzed for each analyte at the specified frequency
       and at the appropriate concentration. Verify that acceptable %R results were obtained.
    4.  Recalculate one or more of the ICV and CCV %R(s) using the following equation and verify that
       the recalculated value agrees with the laboratory-reported values on Form 2-IN.
                                          Found (value)
                                    %R =	 x 100
                                          True (value)
       Where,
       „    , /  i  x      Concentration (in ug/L) of each analyte measured in the analysis of the
       Found (value)   =  T™,   „„,,  \  ..                 J                      J
                          ICV or CCV solution
       True (value)    =  Concentration (in (ig/L) of each analyte in the ICV or CCV source
E.  Action

    NOTES:   For initial calibrations or ICV standards that do not meet the technical criteria,  apply the
               action to all associated samples reported from the analytical sequence.
               For CCV standards that do not meet the technical criteria, apply the action to all samples
               analyzed between a previous technically acceptable analysis of the QC sample and a
               subsequent technically acceptable analysis  of the QC sample in the analytical sequence.
    1.  If the instrument was not calibrated each time the instrument was set up, qualify detects and non-
       detects as unusable (R).  If the instrument was not calibrated with at least the minimum number of
       standards, or if the calibration curve does not include standards at required concentrations (e.g., a
       blank and at least one at or below CRQL but above the MDL), use professional judgment to
       qualify detects as estimated (J) or unusable (R), and non-detects as estimated (UJ) or unusable
       (R).
    2.  If the correlation coefficient is < 0.995, the %Ds are outside the ±30% limit, or the y-intercept >
       CRQL, qualify detects as estimated (J) and non-detects as estimated (UJ).
    3.  If the ICV or CCV %R falls outside the acceptance windows, use professional judgment to
       qualify all associated data. If possible, indicate the bias in the review. The following guidelines
       are recommended:
       a.  If the  ICV or CCV %R is < 75%, use professional judgment to qualify detects as estimated
           low (J-) or unusable (R), and non-detects as unusable (R).

August 2014                                                                                    18

-------
Inorganic Data Review
ICP-AES
       b.  If the ICV or CCV %R falls within the range of 75-89%, qualify detects as estimated low (J-)
           and non-detects as estimated (UJ).
       c.  If the ICV or CCV %R falls within the range of 90-110%, detects and non-detects should not
           be qualified.
       d.  If the ICV or CCV %R falls within the range of 111-125%, qualify detects as estimated high
           (J+). Non-detects should not be qualified.
       e.  If the ICV or CCV %R is > 125%, use professional judgment to qualify detects as estimated
           high (J+) or unusable (R).  Non-detects should not be qualified.
    4.  If the laboratory failed to provide adequate calibration information, notify the Regional
       Laboratory COR. The Regional Laboratory COR may contact the laboratory to request the
       necessary information. If the information is unavailable, use professional judgment to assess the
       data.
    5.  Annotate the potential effects on the reported data due to exceeding the calibration criteria in the
       Data Review Narrative.
    6.  If calibration criteria are grossly exceeded, note this for Regional Laboratory COR action.
    NOTE:     For truly critical samples, a further in-depth evaluation of the calibration curve may be
               warranted to determine if additional qualification is necessary.

                       Table 4.  Calibration Actions for ICP-AES Analysis
Criteria
Calibration not performed
Calibration incomplete
Correlation coefficient < 0.995;
%D outside ±30%; y-intercept > CRQL
ICV/CCV%R<75%
ICV/CCV %R 75-89%
ICV/CCV%R90-110%
ICV/CCV %R 1 1 1-125%
ICV/CCV %R> 125%
Action
Detect
R
Use professional
judgment
JorR
J
Use professional
judgment
J-orR
J-
No qualification
J+
Use professional
judgment
J+orR
Non-detect
R
Use professional
judgment
UJorR
UJ
Use professional
judgment
R
UJ
No qualification
No qualification
No qualification
August 2014
      19

-------
Inorganic Data Review                                                                   ICP-AES

                                         III.    Blanks
A.  Review Items
    Form 1-IN, Form 3-IN, Form 12-IN,  preparation logs, calibration standard logs, instrument logs, and
    raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the blank responses by
    determining the existence and magnitude of contamination resulting from laboratory (or field)
    activities or baseline drift during analysis.

C.  Criteria
    1.  No contaminants should be found in the blank(s).
    2.  The Initial  Calibration Blank (ICB) shall be analyzed at each mass used for analysis after the
       analytical standards, but not before analysis of the ICV during the initial calibration of the
       instrument (see Section II.C.l).
    3.  A Continuing Calibration Blank (CCB) shall be analyzed at each wavelength used for the
       analysis, immediately after every CCV. The CCB shall be analyzed at a frequency of every two
       hours during the analytical sequence. The CCB shall be analyzed at the beginning of the
       analytical sequence, and again after the last CCV that was analyzed after the last analytical
       sample of the analytical sequence. The CCB result (absolute value) shall not exceed the CRQL
       of each analyte for which analysis is performed.
    4.  At least one Preparation Blank shall  be prepared and analyzed for each matrix, with every SDG,
       or with each batch of samples digested, whichever is more frequent.  The Preparation Blank
       consists of reagent water or a clean wipe processed through the appropriate sample preparation
       and analysis procedure.
    5.  If the concentration of any analyte in the Preparation Blank is > CRQL, the lowest concentration
       of that analyte in the associated samples must be > lOx the Preparation Blank concentration.
       Otherwise, all associated samples with the analyte's concentration <  lOx the Preparation Blank
       concentration and > CRQL, should be redigested and reanalyzed for that analyte. The laboratory
       is not to correct the sample concentration for the blank value.
    6.  If the concentration of any analyte in the Preparation Blank is < (-CRQL), all associated samples
       with the analyte's concentration < lOx the CRQL should be redigested and reanalyzed.
    7.  At least one Leachate Extraction Blank (LEB) shall be prepared and analyzed for each batch of
       samples extracted by TCLP or SPLP. The LEB consists of reagent water processed through the
       extraction procedure. Post-extraction, the LEB shall be processed through the appropriate sample
       preparation and analysis procedure.
D.  Evaluation
    1.  Verify that an ICB was analyzed after the calibration, the CCB was analyzed at the specified
       frequency and sequence  during the analysis, and Preparation Blanks were prepared and analyzed
       as appropriate for the SDG (e.g., total number of samples, various types of matrices present,
       number of digestion batches, etc.).

    2.  Review the results reported on Form 3-IN, as well as the raw data for all blanks, and verify that
       the results were accurately reported.
    3.  Evaluate all of the associated blanks for the presence of target analytes. Verify that if the
       concentration of any target analyte was > CRQL in a Preparation Blank, all associated samples
       with the analyte's concentration > CRQL but < lOx the Preparation Blank concentration were
       redigested and reanalyzed for that analyte. Verify that if a concentration was < (-CRQL) in a
       Preparation Blank, all associated samples with the analyte's concentration < lOx CRQL were

August 2014                                                                                   20

-------
Inorganic Data Review                                                                  ICP-AES

       redigested and reanalyzed. Verify that if the absolute value of any target analytes was > CRQL in
       an ICB or a CCB, the analysis was terminated, the problem corrected, the instrument recalibrated,
       and the preceding 10 analytical samples or all analytical samples analyzed since the last
       compliant calibration blank reanalyzed.
E.  Action
    NOTES:  For ICBs that do not meet the technical criteria, apply the action to all associated samples
              reported from the analytical sequence.
              For CCBs that do not meet the technical criteria, apply the action to all associated
              samples analyzed between a previous technically acceptable analysis of the CCB and a
              subsequent technically acceptable analysis of the CCB in the analytical sequence.
              For Preparation Blanks that do not meet the technical criteria, apply the action to all
              associated samples prepared in the same preparation batch. For LEBs that  do not meet
              the technical criteria, apply the action to all associated samples extracted in the  same
              extraction batch.
    1.  If the appropriate blanks were not analyzed with the specified frequency, use professional
       judgment to determine if the associated sample data should be qualified; obtain additional
       information from the laboratory, if necessary.  Record the situation in the Data Review Narrative,
       and note  it for Regional Laboratory COR action.
    2.  Action regarding unsuitable blank results depends on the circumstances and origin  of the blank.
       In instances where more than one blank is associated with a given sample, qualification should be
       based upon a comparison with the associated blank having the highest concentration of
       contaminant.
    3.  Some general "technical" review actions include:
       a.   For any blank (including Preparation Blanks and LEBs) reported with detects < CRQLs,
           report detects < CRQLs at the CRQLs and qualify as non-detect (U). For any blank
           (including Preparation Blanks and LEBs)  reported with detects < CRQLs, use professional
          judgment to qualify the sample results > CRQLs. Non-detects should not be qualified.
       b.   For any blank (including Preparation Blanks and LEBs) reported with a negative result,
           < (-MDL) but > (-CRQL), carefully evaluate and determine its effect on the sample data. Use
           professional judgment to assess the data.
       c.   The blank analyses may not involve the same weights, volumes, or dilution factors as the
           associated samples. In particular, soil/sediment sample results reported on Form 1-IN will
           not be on the same basis (units, dilution) as the calibration blank data reported on Form 3-IN.
           It may be easier to work with the raw data and/or convert the ICB or CCB results to the same
           units as the soil/sediment samples for comparison purposes.
    4.  Specific "method" actions include:
       a.   If an ICB or a CCB result is > CRQL, the analysis should be terminated.  If the analysis was
           not terminated and the associated samples were not reanalyzed, non-detects should not be
           qualified. Report detects < CRQLs at the  CRQLs and qualify as non-detect (U). Report
           sample results that are > CRQLs but < ICB/CCB Results at ICB/CCB Results and use
           professional judgment to qualify as non-detect (U) or unusable (R).  Use professional
          judgment to qualify sample results > ICB/CCB Results.  Record the situation in the Data
           Review Narrative, and note it for Regional Laboratory COR action.
       b.   If an ICB or a CCB result is < (-CRQL), the analysis should be terminated.  If the analysis
           was not terminated and the associated samples were not reanalyzed, use professional
          judgment to qualify non-detects as estimated (UJ) or unusable (R). Use professional
          judgment to qualify detects < CRQL, or qualify as estimated low (J-). Use professional
          judgment to qualify sample results that are > CRQLs as estimated low (J-).

August 2014                                                                                   21

-------
Inorganic Data Review
ICP-AES
       c.  If the concentration of any analyte in the Preparation Blank/LEB is > CRQL, the lowest
           concentration of that analyte in the associated samples must be > lOx the Preparation
           Blank/LEB concentration. All samples associated with the Preparation Blank with
           concentrations < 1 Ox the Preparation Blank concentration and > CRQL should have been
           redigested and reanalyzed. If the associated samples were not redigested and reanalyzed,
           report the sample results at Preparation Blank Results;  use professional judgment to qualify
           the results as estimated high (J+) or unusable (R). Report results < lOx the LEB
           concentration and > CRQL in the samples associated with the LEB at LEB Results; use
           professional judgment to qualify the results as estimated high (J+) or unusable (R).  Report
           detects < CRQLs in the  samples associated with the Preparation Blank/LEB at CRQLs and
           qualify as non-detect (U).  Non-detects and sample results that are > lOx the Preparation
           Blank/LEB Results should not be qualified.  If the laboratory failed to redigest and reanalyze
           the samples associated with the Preparation Blank, record it in the Data Review Narrative,
           and note it for Regional Laboratory COR action.
       d.  For any Preparation Blank or LEB reported with a negative result, < (-CRQL), use
           professional judgment to qualify detects < CRQL, or qualify as estimated low (J-). Qualify
           sample results that are > CRQLs as estimated low (J-), and qualify non-detects as estimated
           (UJ).  Sample results that are > lOx CRQLs should not be qualified.

                          Table 5. Blank Actions for ICP-AES Analysis
Blank
Type
ICB/CCB
ICB/CCB
ICB/CCB
ICB/CCB
Blank Result
Detect < CRQL
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Sample Result
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but
< ICB/CCB Result
> ICB/CCB Result
Non-detect
Detect < CRQL
>CRQL
Action
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at ICB/CCB Result and
qualify as non-detect (U) or unusable
(R)
Use professional judgment
Use professional judgment to qualify
as estimated (UJ) or unusable (R)
Use professional judgment or (J-)
Use professional judgment to qualify
as estimated low (J-)
August 2014
      22

-------
Inorganic Data Review
ICP-AES
Blank
Type
Preparation
Blank/LEB
Preparation
Blank/LEB
Preparation
Blank/LEB
Preparation
Blank/LEB
Blank Result
Detect < CRQL
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Sample Result
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but
< 1 Ox the Preparation
Blank/LEB Result
> lOx the Preparation
Blank/LEB Result
Non-detect
Detect < CRQL
< lOx CRQL
> lOx CRQL
Action
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
No qualification
Report at CRQL and qualify as a non-
detect (U)
Report at Preparation Blank/LEB
Result and use professional judgment
to qualify results as estimated high
(J+) or unusable (R)
No qualification
Qualify as estimated (UJ)
Use professional judgment or (J-)
Qualify results that are > CRQL as
estimated low (J-)
No qualification
August 2014
      23

-------
Inorganic Data Review                                                                    ICP-AES

                               IV.    Interference Check Sample
A.  Review Items
    Form 4-IN, Form 12-IN, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the instrument's ability to
    overcome interferences typical of those found in samples.

C.  Criteria
    1.  The Interference Check Sample (ICS) consists of two solutions: Solution A and Solution AB.
       Solution A consists of the interferents, and Solution AB consists of the analytes mixed with the
       interferents. An ICS analysis consists of analyzing both solutions consecutively, starting with
       Solution A, for all wavelengths used for each analyte reported by ICP-AES.
    2.  An ICS must be analyzed undiluted at the beginning of each sample analysis sequence. The ICS
       is not to be analyzed prior to the ICV, and shall be immediately followed by a CCV, followed by
       aCCB.

    3.  Results for the analysis of ICS  Solution A must fall within the control limits of ± CRQL or ± 20%
       of the true value (whichever is  greater) for the analytes and interferents included in the solution.
    4.  Results for the analysis of ICS  Solution AB must fall within the control limits of ± CRQL or ±
       20% of the true value  (whichever is greater) for the analytes and interferents included in the
       solution.
    5.  If the value of an ICS  result exceeds ± 2x the CRQL, or ± 20% of true value (whichever is
       greater) criteria, the analysis shall be terminated, the problem corrected, the instrument
       recalibrated, the new calibration then reverified, and all analytical samples since the last
       compliant ICS reanalyzed.
    6.  The ICS should be obtained from the EPA, if available, and analyzed according to the
       instructions supplied with the solutions. If the ICS is not available from the EPA, an independent
       ICS solution shall be prepared using certified standards with the interferent and analyte
       concentrations at the levels specified in the method.
D.  Evaluation
    1.  Verify, using Form 12-IN and the raw data [Inductively Coupled Plasma (ICP) instrument
       printout], that the ICS was analyzed at the specified frequency and sequence during the analytical
       sequence.
    2.  Evaluate the ICS raw data for results with an absolute value that is > MDL for those analytes that
       are not present in the ICS solution.
    3.  Recalculate, using the raw data and the following equation, one or more of the analyte %Rs, and
       verify that the recalculated value agrees with the laboratory-reported values on Form 4-IN.
                                         Found (value)
                                    %R=	 x 100
                                          True (value)

       Where,
        F    , ,  ,   ,    _ Concentration (in (ig/L) of each analyte interferent measured in the
              (value;    - walysis of IC§ Solution A or ICS Solution AB

        „    ,   ,   ,     _ Concentration (in (ig/L) of each analyte or interferent in ICS Solution
        1 rue (value)     - A or ICS Solutlon AB
August 2014                                                                                   24

-------
Inorganic Data Review                                                                    ICP-AES

    4.  If the value of an ICS result exceeds the ± CRQL or ± 20% of true value (whichever is greater)
       criteria, and the laboratory failed to terminate the analysis and take the appropriate corrective
       action, note this for Regional Laboratory COR action and record the situation in the Data Review
       Narrative.  Use professional judgment to assess the data.
E.  Action
    NOTE:    For an ICS that does not meet the technical criteria, apply the action to all samples
               reported from the analytical sequence.
    1.  If the ICS was not analyzed at the specified frequency, qualify detects and non-detects as
       unusable (R).  If the ICS was analyzed, but not in the proper sequence, use professional judgment
       to qualify detects and non-detects.
    2.  The raw data may not contain results for interferents.  In this case, use professional judgment to
       qualify the data.  If the data contains results for interferents, apply the following actions to
       samples with concentrations of interferents that are within 10% of the levels of the interferents in
       the ICS:
       a.  If the ICS Solution AB %R for an analyte or interferent is < 50%, qualify detects as estimated
           low (J-) and non-detects as unusable (R).
       b.  If the ICS %R for an analyte or interferent falls within the range of 50-79% [or the ICS found
           value is < (true value - CRQL), whichever is lower], qualify detects as estimated low (J-), and
           non-detects as estimated (UJ).
       c.  If the ICS %R for an analyte or interferent falls within the range of 80-120%, detects and
           non-detects should not be qualified.
       d.  If the ICS %R for an analyte or interferent is > 120% [or the ICS found value is > (true value
           + CRQL), whichever is greater], qualify detects as estimated high (J+).  Non-detects should
           not be qualified.
       e.  If the ICS %R for an analyte or interferent is above 150%, use professional judgment to
           determine the qualifications of the associated sample data.
    3.  If sample results that are > MDLs are observed for analytes that are not present in the ICS
       solution, the possibility of false positives exists. An evaluation of the associated sample data for
       the affected analytes should be made.  For samples with comparable or higher levels of
       interferents and with analyte concentrations that approximate those levels found in the ICS,
       qualify detects as estimated high (J+). Non-detects should not be qualified.
    4.  If negative sample results are observed for analytes that are not present in the ICS solution, and
       their absolute values are > MDLs, the possibility of false negatives in the samples exists. An
       evaluation of the  associated sample data for the affected analytes should be made. For samples
       with levels of interferents that are comparable to or higher than the levels found in the ICS,
       qualify detects <  lOx the absolute value of the negative result as estimated low (J-), and qualify
       non-detects as estimated (UJ).
    NOTE:    The same result units should be used when comparing analyte results in samples to those
               in the ICS.  Unit conversion may be necessary when soil/sediment or wipe samples are
               evaluated.
    5.  In general, ICP-AES sample data can be accepted if the concentrations of Aluminum (Al),
       Calcium (Ca), Iron (Fe), and Magnesium (Mg) in the sample are found to be less than or equal to
       their respective concentrations in the ICS. If these elements are present at concentrations greater
       than the level in the ICS,  or other elements are present in the sample at > 10 mg/L, investigate the
       possibility of other interference effects as given in the ICP-AES method  or as indicated by the
       laboratory's interelement correction factors reported on Forms 10A-IN and  10B-IN for that
       particular instrument. The analyte concentration equivalents presented in the method should be
       considered only as estimated values since the exact value of any analytical system is instrument-
August 2014                                                                                   25

-------
Inorganic Data Review
ICP-AES
       specific. Therefore, estimate the concentration produced by an interfering element.  If the
       estimate is > 2xthe CRQL, and also > 10% of the reported concentration of the affected element,
       qualify the affected results as estimated (J).
    6.  If the raw data does not contain results for the interferents, annotate this in the Data Review
       Narrative.
    7.  Actions regarding the interpretation and/or the subsequent qualification of ICP data due to the
       ICS analytical results can be extremely complex.  Use professional judgment to determine the
       need for the associated sample data to be qualified; obtain additional information from the
       laboratory, if necessary. Record all interpretive situations in the Data Review Narrative.
    8.  If the ICS acceptance criteria are grossly exceeded, note the specifics for Regional Laboratory
       COR action.

                   Table 6.  Interference Check Actions for ICP-AES Analysis
Criteria
ICS not analyzed
ICS not analyzed in the proper sequence
ICSAB%R<50%
ICS %R 50-79% [or ICS found value is < (true
value - CRQL) whichever is lower]
ICS %R 80-120%
ICS %R > 120% (or ICS found value is > (true
value + CRQL) whichever is greater]
ICS%R>150%
Sample results > MDLs, but not present in ICS
Negative sample results, but not present in ICS
Action
Detect
R
Use professional
judgment
J-
J-
No qualification
J+
Use professional
judgment
J+
J-
for results <
lOx (negative sample
result)
Non- detect
R
Use professional
judgment
R
UJ
No qualification
No qualification
Use professional
judgment
No qualification
UJ
August 2014
      26

-------
Inorganic Data Review                                                                  ICP-AES

                              V.      Laboratory Control Sample
A.  Review Items
    Form 7-IN, preparation logs, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the recovery of the
    digested Laboratory Control Sample (LCS).

C.  Criteria
    1.  Aqueous/water, soil/sediment and wipe LCSs shall be analyzed for each analyte utilizing the
       same sample preparations, analytical methods, and Quality Assurance/Quality Control (QA/QC)
       procedures as employed for the samples.
       a.   One LCS shall be prepared and analyzed for every group of aqueous/water or soil/sediment
           samples in an SDG, or with each batch of samples digested, whichever is more frequent. The
           LCS shall be spiked such that the final digestate contains each analyte at 2x the CRQL for the
           associated matrix.
       b.   One LCS shall be prepared and analyzed for each group of wipe samples in an SDG, or with
           each batch of wipe  samples digested, whichever is more frequent. The wipe LCS shall be
           spiked such that the final digestate contains each analyte at 2x the CRQL for the associated
           matrix.

       c.   All LCS %Rs must fall within the control limits of 70-130%, except for Antimony (Sb) and
           Silver (Ag) which must fall within the control limits of 50-150%.  If the %R for the
           aqueous/water and soil/sediment LCS falls outside of the control limits, the analysis should
           be terminated, the problem corrected, and the samples prepared with that LCS redigested and
           reanalyzed. No corrective actions are required for wipe LCSs when the %R is outside the
           control limits.
D.  Evaluation
    1.  Verify, using Form 7-IN, preparation logs, and raw data, that the appropriate number of required
       LCSs were prepared and analyzed for the SDG.
    2.  Evaluate Form 7-IN and verify that all results for each analyte fall within the established control
       limits.
       a.   Check the raw data to verify that the %Rs on Form 7-IN were accurately transcribed.
           Recalculate one or more of the reported %Rs using the following equation:

                                         Found (value)
                                    %R=-	-^-T xlOO
                                         I rue (value)

       Where,
       „     1,1^      Concentration of each analyte (in ug/L, mg/kg, or ug) measured in the
       Found (value)   =     ,  .   ri,  T „„      j   v   n&  ,   &  &,   \*&>
                          analysis or the LCS
       True (value)     =  Concentration of each analyte (in (ig/L, mg/kg, or ug) in the LCS

    3.  Verify that the LCS was prepared at the same time as the associated samples using the same
       procedures.
August 2014                                                                                 27

-------
Inorganic Data Review                                                                   ICP-AES

E.  Action
    NOTE:     If the LCS criteria are not met, the laboratory performance and method accuracy are in
               question. Use professional judgment to determine if the data should be qualified or
               rejected. The following guidance is suggested for qualifying sample data associated with
               an LCS that does not meet the required criteria.
               For an LCS that does not meet the technical criteria, apply the action to all samples in the
               same preparation batch.
    1.  If the required LCS was not analyzed at the specified frequency, use professional judgment to
       determine if the associated sample results should be qualified; obtain additional information from
       the laboratory, if necessary. If a laboratory fails to analyze a LCS with each SDG, or if a
       laboratory consistently fails to generate acceptable LCS recoveries, record the  situation in the
       Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated  (UJ).
    2.  Aqueous/Water and Soil/Sediment LCS:
       a.   If LCS  %R is < 40% (< 20% for Ag and  Sb), qualify detects as estimated low (J-) and non-
           detects as unusable (R).
       b.   If the LCS %R falls within the range of 40-69% (20-49% for Ag and Sb), qualify detects as
           estimated low (J-) and non-detects as estimated (UJ).
       c.   If the LCS %R falls within the range of 70-130%, detects and non-detects  should not be
           qualified.
       d.   If the LCS %R is > 130% (150% for Ag and Sb), qualify detects as estimated high (J+).
           Non-detects  should not be qualified.
       e.   If the LCS %R is > 150% (170% for Ag and Sb), qualify detects as unusable (R). Non-
           detects should not be qualified.
    3.  Wipe LCS:
       a.   If the LCS %R is < 40%, qualify detects  as estimated low (J-) and non-detects as unusable
           (R).
       b.   If the LCS %R is in the range of 40-69%, qualify detects as estimated low  (J-) and non-
           detects as estimated (UJ).
       c.   If the LCS %R is within 70-130%, detects and non-detects should not be qualified.
       d.   If the LCS %R is > 130%, qualify detects as estimated high (J+). Non-detects should not be
           qualified.
    4.  Annotate the potential effects on the data due to out-of-control LCS results in the Data Review
       Narrative.
August 2014                                                                                  28

-------
Inorganic Data Review
ICP-AES
                        Table 7. LCS Actions for ICP-AES Analysis
Criteria
Aqueous/Water and Soil/Sediment %R < 40%
(< 20% Ag, Sb)
Aqueous/Water and Soil/Sediment %R 40-69%
(20-69% Ag, Sb)
Aqueous/Water and Soil/Sediment %R 70-130%
Aqueous/Water and Soil/Sediment
%R > 130% (150% Ag, Sb)
Aqueous/Water and Soil/Sediment
%R > 150% (170% Ag, Sb)
Wipe%R<40%
Wipe %R 40-69%
Wipe %R 70- 130%
Wipe %R> 130%
Action
Detect
J-
J-
No qualification
J+
R
J-
J-
No qualification
J+
Non-detect
R
UJ
No qualification
No qualification
No qualification
R
UJ
No qualification
No qualification
August 2014
      29

-------
Inorganic Data Review                                                                  ICP-AES

                               VI.    Duplicate Sample Analysis
A.  Review Items
    Cover Page, Form 6-IN, instrument printouts, and raw data.

B.  Objective
    The objective of duplicate sample analysis is to demonstrate acceptable method precision by the
    laboratory at the time of analysis.

C.  Criteria
    1 .   Samples identified as field blanks or Performance Evaluation (PE) samples cannot be used for
        duplicate sample analysis.
    2.   At least one duplicate sample shall be prepared and analyzed from each group of samples of a
        similar matrix type (e.g., water or soil) or for each SDG. Duplicates are not required for wipe
        samples. Duplicates cannot be averaged for reporting on Form 1-IN. Additional duplicate
        sample analyses may be required by EPA Regional request. Alternately, the Region may require
        that a specific sample be used for the duplicate  sample analysis.
    3 .   A control limit of 20% for the Relative Percent Difference (RPD) shall be used for original and
        duplicate sample values > 5x the CRQL.
    4.   A control limit of the CRQL shall be used if either the sample or duplicate value is < 5x the
        CRQL.  The absolute value of the control limit (CRQL)  shall be entered in the "Control Limit"
        column on Form 6-IN. If both samples are non-detects, the RPD is not calculated for Form 6-IN.
    NOTE:     The above control limits are method requirements for duplicate samples, regardless of
               the sample matrix type. However, it should be noted that laboratory variability arising
               from the sub-sampling of non-homogenous soil  samples is a common occurrence.
               Therefore, for technical review purposes only, Regional policy or project DQOs may
               allow the use of less restrictive criteria  (e.g.,  35% RPD, 2x the  CRQL) to be assessed
               against duplicate soil samples.
D.  Evaluation
    1 .   Verify, using the Cover Page and the raw data,  that the appropriate number of required duplicate
        samples were prepared and analyzed for the SDG.
    2.   Verify, using Form 6-IN and the raw data, that  all duplicate results for each analyte fall within the
        established control limits.
    3.   Verify that a field blank or PE sample was not used for duplicate analysis.
    4.   Check the raw data and recalculate one or more of the RPD values using the following equation
        to verify that the results were correctly reported on Form 6-IN:

                                             IS-DI
       Where,
       RPD   =  Relative Percent Difference
       S      =  Sample Result (original)
       D      =  Duplicate Result
August 2014                                                                                 30

-------
Inorganic Data Review                                                                   ICP-AES

E.  Action
    NOTE:     For a duplicate sample analysis that does not meet the technical criteria, apply the action
               to all samples of the same matrix if the samples are considered sufficiently similar.
               Exercise professional judgment in determining sample similarity when making use of all
               available data, including: site and sampling documentation (e.g., location and type of
               sample, descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity,
               chlorine); and laboratory data for other parameters [e.g., Total Suspended Solids (TSS),
               Total Dissolved Solids (TDS), Total Organic Carbon (TOC), alkalinity or buffering
               capacity, reactive sulfide, anions].  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG. Two
               determinations are: 1) only some samples in the SDG are similar to the duplicate  sample,
               and that only these samples should be qualified; or 2) no samples are sufficiently similar
               to the sample used for the duplicate, and thus only the field sample used to prepare the
               duplicate sample should be qualified.
    1.  If the appropriate number of duplicate samples was not analyzed for each matrix using the correct
       frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary.  Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If both original sample and duplicate sample results are > 5x the CRQL and the RPD is > 20%,
       qualify detects as estimated (J) and non-detects as estimated (UJ).
    3.  If RPD > 100%, use professional judgment to determine if the associated  sample data should be
       qualified.
    4.  If both original sample and duplicate sample results are > 5x the CRQL and the RPD is < 20%,
       detects and non-detects should not be qualified.
    5.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate > CRQL, qualify detects as estimated (J) and
       non-detects as estimated (UJ).
    6.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate < CRQL, detects and non-detects should not be
       qualified.
    7.  If a field blank or PE sample was used for the duplicate sample  analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be  carefully checked. Use
       professional judgment when evaluating the data.
    8.  Annotate the potential effects on the  data due to out-of-control duplicate sample results in the
       Data Review Narrative.
August 2014                                                                                  31

-------
Inorganic Data Review
ICP-AES
                    Table 8. Duplicate Sample Actions for ICP-AES Analysis
Criteria
Both original sample and duplicate sample results are > 5x
the CRQL and RPD > 20%*
RPD > 100%
Both original sample and duplicate sample results are > 5x
the CRQL and RPD is < 20%
Original sample or duplicate sample result < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate > CRQL*
Original sample or duplicate sample result < 5x the
CRQL (including non-detects) and absolute difference
between sample and duplicate < CRQL
Action
Detect
J
Use professional
judgment
No qualification
J
No qualification
Non-detect
UJ
Use professional
judgment
No qualification
UJ
No qualification
    *   The above control limits are method requirements for duplicate samples, regardless of the
       sample matrix type. However, it should be noted that laboratory variability arising from the sub-
       sampling of non-homogenous soil samples is a common occurrence.  Therefore, for technical
       review purposes only, Regional policy or project DQOs may allow the use of less restrictive
       criteria (e.g., 35% RPD, 2xthe CRQL) to be assessed against duplicate soil samples.
August 2014
      32

-------
Inorganic Data Review                                                                   ICP-AES

                                 VII.   Spike Sample Analysis
A.  Review Items
    Cover Page, Form 5A-IN, Form 5B-IN, instrument printouts, and raw data.

B.  Objective
    The objective of the spiked sample analysis is to evaluate the effect of each sample matrix on the
    sample preparation procedures and the measurement methodology.

C.  Criteria
    1.   Samples identified as field blanks or PE samples cannot be used for spiked sample analysis.
    2.   At least one spiked sample (pre-digestion) shall be prepared and analyzed from each group of
        samples with a similar matrix type (e.g., water or soil), or for each SDG.  Matrix Spikes are not
        required for wipe samples.
    3.   When the Matrix Spike recovery falls outside of the control limits and the sample result is < 4x
        the spike added, a post-digestion spike shall be performed for those analytes that do not meet the
        specified criteria. An aliquot of the remaining unspiked sample shall be spiked at 2x the
        indigenous level or 2x the CRQL, whichever is greater.
    NOTE:     Post-digestion spikes are not required for Ag.
    4.   The spike %R shall be within the established acceptance limits. However, spike recovery limits
        do not apply when the sample concentration is > 4x the spike added. In such an event, the data
        shall be reported unflagged, even if the %R does not meet the acceptance criteria.
    5.   If the spiked sample analysis was performed on the same sample that was chosen for the duplicate
        sample analysis, spike calculations shall be performed using the results of the sample designated
        as the "original sample." The average of the duplicate  results cannot be used for the purpose of
        determining %R.
    NOTE:     The final spike  concentrations required for the  various target analytes are presented in the
               methods described in the SOW.
D.  Evaluation
    1.   Verify, using the Cover Page, Form 5 A-IN, and raw data, that the appropriate number of required
        spiked samples was prepared and analyzed for the  SDG.
    2.   Verify that a field blank or PE sample was not used for the spiked sample analysis.
    3.   Verify, using Form 5A-IN, and the raw data, that all pre-digestion spiked sample results  for each
        required analyte fall within the established control limits.  If not, verify that a post-digestion spike
        was prepared and analyzed.
    4.   Recalculate, using the raw data,  one or more of the %R using the following equation, and verify
        that the recalculated value agrees with the laboratory-reported values on Forms 5A-IN & 5B-IN:

                                                SSR-SR
                                  %Recovery  = ——— x 100
                                                  SA

        Where,
        SSR    =   Spiked Sample Result
        SR     =   Sample Result
        SA     =   Spike Added
August 2014                                                                                  33

-------
Inorganic Data Review                                                                    ICP-AES

    NOTE:     When the sample result is < MDL or reported as a non-detect, use SR = 0 only for the
               purpose of calculating the %R. The actual spiked sample results, sample results, and %R
               (positive or negative) shall still be reported on Forms 5A-IN & 5B-IN.
E.  Action
    NOTE:     For a Matrix Spike that does not meet the technical criteria, apply the action to all
               samples of the same matrix if the samples are considered sufficiently similar.  Exercise
               professional judgment in determining sample similarity when making use of all available
               data, including:  site and sampling documentation (e.g., location and type of sample,
               descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity, chlorine);
               and laboratory data for other parameters (e.g., TSS, TDS, TOC, alkalinity or buffering
               capacity, reactive sulfide, anions). Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG. Two
               determinations are: 1) only some of the samples in the SDG are similar to the Matrix
               Spike sample, and that only these samples should be qualified; or 2) no samples are
               sufficiently similar to the sample used for the Matrix Spike, and thus only the field
               sample used to prepare the Matrix Spike sample should be qualified.
    1.  If the appropriate number of Matrix Spike samples was not analyzed for each matrix using the
       correct frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary. Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If a field blank or PE sample was used for the spiked sample analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be carefully checked.  Use
       professional judgment when evaluating the data. Detects should be qualified as estimated (J) and
       non-detects qualified as estimated (UJ).
    3.  If the Matrix Spike recovery does not meet the evaluation criteria and a required post-digestion
       spike was not performed, note this for Regional Laboratory COR action.
    4.  If the Matrix Spike %R is < 30%, verify that a post-digestion spike was analyzed. If the post-
       digestion spike %R is < 75% or the analysis was not performed, qualify detects as estimated low
       (J-), and qualify non-detects as unusable (R). If the post-digestion spike %R is > 75%, qualify
       detects as estimated (J) and qualify non-detects as estimated (UJ).
    5.  If the Matrix Spike %R falls within the range of 30-74%, verify that a post-digestion spike was
       analyzed (if required when sample concentration is < 4x spike added). If the post-digestion spike
       %R < 75% or the analysis  was not performed, qualify detects as estimated low (J-) and non-
       detects as estimated (UJ).  If the post-digestion spike %R> 75%, qualify detects as estimated (J)
       and non-detects as estimated (UJ).
    6.  If the Matrix Spike %R falls within the range of 75-125%, no post-digestion spike is required.
       Detects and non-detects should not be qualified.
    7.  If the Matrix Spike %R is > 125%, verify that a post-digestion spike was analyzed (if required
       when sample concentration is < 4x spike added). If the post-digestion spike %R > 125% or the
       analysis was not performed, qualify detects as estimated high (J+); non-detects should not be
       qualified. If the post-digestion spike %R< 125%,  qualify detects as estimated (J); non-detects
       should not be qualified.
    8.  Annotate the potential effects on the data due to out-of-control spiked sample results in the Data
       Review Narrative.
August 2014                                                                                   34

-------
Inorganic Data Review
ICP-AES
                      Table 9. Spike Sample Actions for ICP-AES Analysis
Criteria
Matrix Spike %R < 30%
Post-digestion spike %R < 75%
Matrix Spike %R < 30%
Post-digestion spike %R> 75%
Matrix Spike %R 30-74%
Post-digestion Spike %R < 75%
Matrix Spike %R 30-74%
Post-digestion spike %R> 75%
Matrix Spike %R> 125%
Post-digestion spike %R> 125%
Matrix Spike %R> 125%
Post-digestion spike %R< 125%
Matrix Spike %R < 30%
No post-digestion spike performed
(e.g., not required for Ag)
Matrix Spike %R 30-74%
No post-digestion spike performed
(e.g., not required for Ag)
Matrix Spike %R 75-125%
Post-digestion spike not required
Matrix Spike %R> 125%
No post-digestion spike performed
(e.g., not required for Ag)
Action
Detect
J-
J
J-
J
J+
J
J-
J-
No qualification
J+
Non- detect
R
UJ
UJ
UJ
No qualification
No qualification
R
UJ
No qualification
No qualification
    NOTE:    The above control limits are method requirements for spike samples, regardless of the
              sample matrix type.  However, it should be noted that laboratory variability arising from
              the sub-sampling of non-homogenous soil samples is a common occurrence. Therefore,
              for technical review purposes only, Regional policy or project DQOs may allow the use
              of less restrictive criteria (e.g., 10 %R and 150 %R for the lower and upper limits) to be
              assessed against spike and post-digestion spike soil samples.
August 2014
      35

-------
Inorganic Data Review                                                                    ICP-AES

                                     VIII.   Serial Dilution
A.  Review Items
    Form 1-IN, Form 8 -IN, instrument printouts, and raw data.

B.  Objective
    The objective of the serial dilution analysis is to determine whether or not significant physical or
    chemical interferences exist due to sample matrix.

C.  Criteria
    1 .  An ICP Serial Dilution analysis shall be performed on a sample from each group of samples with
       a similar matrix type (e.g., water, soil or wipe) or for each SDG, whichever is more frequent.
    2.  Samples identified as field blanks or PE samples cannot be used for the ICP Serial Dilution
       analysis.
    3.  If the analyte concentration is sufficiently high (concentration in the original sample is > 5 Ox the
       MDL), the %D between the original determination and the serial dilution analysis (a five-fold
       dilution) after correction for dilution (concentration in the serial dilution sample is > CRQL) shall
       be < 10%.

    NOTE:    The above criteria are method requirements for serial dilution samples, regardless of the
               sample matrix type. However, for technical review purposes only, Regional policy or
               project DQOs may allow the use of less restrictive criteria (e.g., %D > 15%) to be
               assessed against serial dilution soil samples.
D.  Evaluation
    1 .  Verify that a field blank or PE sample was not used for the serial dilution analysis.
    2.  Check the raw data and recalculate the %D using the following equation. Verify that the  serial
       dilution analysis results and the calculated %D results agree with the values reported by the
       laboratory on Form 8 -IN:
                                    %Difference = - - - x 100
        Where,
        I       =   Initial Sample Result
        S      =   Serial Dilution Result

    3.   Check the raw data for any evidence of positive or negative interference (results from the diluted
        sample which are significantly different than the original sample), possibly due to high levels of
        dissolved solids in the sample, ionization effects, etc.
E.  Action
    NOTE:     For a serial dilution that does not meet the technical criteria, apply the action to all
               samples of the same matrix if the samples are considered sufficiently similar.  Exercise
               professional judgment in determining sample similarity when making use of all available
               data, including: site and sampling documentation (e.g., location and type of sample,
               descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity, chlorine);
               and laboratory data for other parameters (e.g., TSS, TDS, TOC, alkalinity or buffering
               capacity,  reactive sulfide, anions).  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG. Two
               determinations are: 1) only some of the samples in the SDG are similar to the serial
               dilution sample, and that only these samples should be qualified; or, 2) no samples are
               sufficiently similar to the sample used for serial dilution, and thus only the field sample
               used to prepare the serial dilution sample should be qualified.

August 2014                                                                                   36

-------
Inorganic Data Review
ICP-AES
    1.  If the appropriate number of serial dilution samples was not analyzed for each matrix, use
       professional judgment to determine if the associated sample data should be qualified; obtain
       additional information from the laboratory, if necessary.  If a field blank or PE sample was used
       for the serial dilution analysis, record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action.  Detects should be qualified as estimated (J) and non-detects as
       estimated (UJ).
    2.  If the analyte concentration in the original sample is > 5 Ox the MDL, its concentration in the
       serial dilution sample is > CRQL, and the %D > 10%, qualify detects as estimated (J) and non-
       detects as estimated (UJ).
    3.  If the analyte concentration in the original sample is > 5 Ox the MDL, its concentration in the
       serial dilution sample is > CRQL, and the %D is < 10%, detects and non-detects should not be
       qualified.
    4.  If the analyte concentration in the original sample is > 5 Ox the MDL, its concentration in the
       serial dilution sample is > CRQL, and %D is > 100%, use professional judgment to determine if
       the associated sample data should be qualified.
    5.  If the analyte concentration in the original sample is > 5x the CRQL and its concentration in the
       serial dilution sample is < CRQL, detects and non-detects should not be qualified.
    6.  If evidence of positive or negative interference is found, use professional judgment to qualify the
       associated sample data. Annotate the potential effects on the reported data in the Data Review
       Narrative.

                     Table 10.  Serial Dilution Actions for ICP-AES Analysis
Criteria
Sample concentration
> 5 Ox MDL, serial dilution sample concentration >
CRQL, and %D > 10%*
Sample concentration
> 5 Ox MDL, serial dilution sample concentration >
CRQL, and %D < 10%
Sample concentration
> 5 Ox MDL, serial dilution sample concentration >
CRQL, and %D is > 100%
Sample concentration > 5x CRQL and serial dilution
sample concentration < CRQL
Interferences present
Action
Detect
J
No qualification
Use professional
judgment
No qualification
Use professional
judgment
Non- detect
UJ
No qualification
Use professional
judgment
No qualification
Use professional
judgment
       The above criteria are method requirements for serial dilution samples, regardless of the sample
       matrix type. However, for technical review purposes only, Regional policy or project DQOs
       may allow the use of less restrictive criteria (e.g., %D > 15%) to be assessed against serial
       dilution soil samples.
August 2014
      37

-------
Inorganic Data Review                                                                  ICP-AES

                    IX.     Regional Quality Assurance and Quality Control
A.  Review Items
    Form 1-IN, instrument printouts, and raw data.

B.  Objective
    The objective is to use results from the analysis of Regional QA/QC samples such as field blanks, PE
    samples, blind spikes, and blind blanks to determine the validity of the analytical results.

C.  Criteria
    Criteria are determined by each Region.
D.  Evaluation
    Evaluation procedures must follow the Region's Standard Operating Procedure (SOP) for data
    review. Each Region will handle the evaluation of PE samples on an individual basis. Compare
    results for PE samples to the acceptance criteria for the specific PE samples if possible.
    Calculate the RPD between field duplicates and provide this information in the Data Review
    Narrative.
E.  Action
    Any action must be  in accordance with Regional specifications and criteria for acceptable PE sample
    results. Note any unacceptable PE sample results for Regional Laboratory COR action.
August 2014                                                                                 38

-------
Inorganic Data Review                                                                    ICP-AES

                               X.      Overall Assessment of Data
A.  Review Items
    Entire sample data package, data review results, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data (including any confirmation data).

B.  Objective
    The objective is to provide the overall assessment on data quality and usability.

C.  Criteria
    1.  Review all available materials to assess the overall quality of the data, keeping in mind the
       additive nature of analytical problems.
    2.  Reported analyte concentrations must be quantitated according to the appropriate analytical
       method, as listed in the method. All sample results must be within the linear calibration ranges
       per methods. Percent Solids (%Solids) must be properly used for all applicable matrix result
       calculations.
D.  Evaluation
    Examine the raw data to verify that correct calculations of the sample results were reported by the
    laboratory. Digestion logs, instrument printouts, etc., should be compared to the reported sample
    results recorded on the appropriate Inorganic Summary Forms (Form 1-IN through Form 16-IN).
    1.  Evaluate any technical problems not previously addressed.
    2.  Examine the raw data for any anomalies (e.g., baseline shifts, negative absorbance, omissions,
       illegibility, etc.).
    3.  Verify that appropriate methods and amounts were used in preparing the samples for analysis.  If
       reduced volumes were used, verify that the laboratory had received Regional Laboratory COR
       approval for the use of the reduced volume.
    4.  Verify that there are no transcription or reduction errors (e.g., dilutions, %Solids, sample weights,
       etc.) on one or more samples. Recalculate %Solids for at least 10% of the samples and verify that
       the calculated %Solids agree with that reported by the laboratory.
    5.  Verify that MDLs are properly reported and that they are not greater than the respective CRQLs.
    6.  Verify that results fall within the calibrated range(s) of the ICP instrument(s) (Form 15-IN).
    7.  If appropriate information is available, assess the usability of the data to assist the data user in
       avoiding inappropriate use of the data.  Review all available information, including the Quality
       Assurance Project Plan (QAPP), focusing specifically on the acceptance or performance criteria,
       the SOPs, and communication with the user concerning the intended use and desired quality of
       these data.
E.  Action
    1.  Use professional judgment to determine if there is any need to qualify data which were not
       qualified based on the QC criteria previously discussed.
    2.  Use professional judgment to qualify detects and non-detects if the MDL exceeds CRQL.
    3.  If a sample is not diluted properly when sample results exceed the upper limit of the calibration
       range, qualify detects as estimated (J).
    4.  Write a brief Data Review Narrative to give the user an indication of the analytical limitations  of
       the data. Annotate any discrepancies between the data and the SDG Narrative for Regional
       Laboratory COR action. If sufficient information on the intended use and required quality of the
       data is available,  include an assessment of the data usability within the given context.
August 2014                                                                                   39

-------
Inorganic Data Review                                                                    ICP-AES

    5.  If any discrepancies are found, notify the Regional Laboratory COR.  The Regional Laboratory
       COR may contact the laboratory to obtain additional information for resolution.  If a discrepancy
       remains unresolved, use professional judgment to determine if qualification of the data is
       warranted.
August 2014                                                                                   40

-------
Inorganic Data Review                                                                  ICP-AES

                                     XI.    Calculations
Aqueous/Water Samples:
    The concentrations determined in the digestate are to be reported in units of (ig/L:
                                                       Vf
                               Concentration (ug/L) = C x — x DF

    Where,

    C      =   Instrument value in |o,g/L (the average of all replicate exposures)
    Vf     =   Final digestion volume (mL)
    V      =   Initial aliquot amount (mL)
    DF    =   Dilution Factor

Soil/Sediment Samples:
    The concentrations determined in the digestate are to be reported on the basis of the dry weight of the
    sample, in units of mg/kg:
                                                            Vf
                      Concentration (mg/kg dry weight) = C x ——- x DF/1000
    Where,
    C      -   Instrument value in |o,g/L (the average of all replicate exposures)
    Vf     =   Final digestion volume (mL)
    W     =   Initial aliquot amount (g)
    S      =   %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    =   Dilution Factor

Wipe Mass:

                                 Mass (ug) = C x Vf x DF/1000

   Where,

   C      =  Instrument value in |o,g/L (The average of all replicate exposures)
   Vf     =  Final digestion volume (mL)
   DF     =  Dilution Factor

Adjusted MDL/Adjusted CRQL Calculation:
    To calculate the adjusted MDL  or adjusted CRQL for aqueous/water samples, substitute the value of
    the MDL ((ig/L) or CRQL  ((ig/L) into the "C" term in the equation above.

    Calculate the adjusted MDL or  adjusted CRQL for soil/sediment samples as follows:
                                                         WM    Vf
                    Adjusted MDL or CRQL (mg/kg) = C x —^-  x —1 x DF
                                                        W x S   VM

   Where,
   C      =  MDL or CRQL (mg/kg)
   WM    =  Minimum method required aliquot amount (g) (1.00 g or 0.50 g)
   W     =  Initial aliquot amount (g)
   VM    =  Method required final sample digestion volume (mL) (100 mL or 50 mL)
   Vf     =  Final digestion volume (mL)
   S      =  % Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
   DF     =  Dilution Factor
August 2014                                                                                 41

-------
Inorganic Data Review                                                               ICP-AES

Hardness (Total) in Aqueous/Water Samples:
   Total Hardness is defined as the sum of calcium and magnesium concentration, expressed as calcium
   carbonate in mg/L.
   Calculate Total Hardness for Aqueous/Water samples as follows:

             Hardness (mg/L) = [Cone. Ca (mg/L) x 2.497] + [Cone. Mg (mg/L) x 4.118]
   Where,
   Cone. Ca (mg/L)    =  Calcium concentration ((ig/L) / 1000
   Cone. Mg (mg/L)   =  Magnesium concentration ((ig/L) / 1000
August 2014                                                                              42

-------
Inorganic Data Review                                                                   ICP-MS

                                  ICP-MS DATA REVIEW
The inorganic data requirements for Inductively Coupled Plasma-Mass Spectrometry (ICP-MS) to be
reviewed during validation are listed below:
Example Analytical Sequence	45
I.    Preservation and Holding Times	47
II.   Tune Analysis	49
III.  Calibration	51
IV.  Blanks	54
V.   Interference Check Sample	58
VI.  Laboratory Control Sample	61
VII.  Duplicate Sample Analysis	63
VIII. Spike Sample Analysis	66
IX.  Serial Dilution	69
X.   Internal Standards	71
XI.  Regional Quality Assurance and Quality Control	73
XII.  Overall  Assessment of Data	74
XIII. Calculations	76
August 2014                                                                                 43

-------
Inorganic Data Review                                                                    ICP-MS
                               This page is intentionally left blank.
August 2014                                                                                  44

-------
Inorganic Data Review                                                            ICP-MS

                              Example Analytical Sequence
The following is an example of an analytical sequence:

Tune
S##
S##
s##
s##
s##
s##
ICV
ICB
ICSA
ICSAB
ccv###
CCB ###
samples
ccv###
CCB###
samples
ccv###
CCB###, etc.
* Suffix ## and ### are as specified in Exhibit B of the Statement of Work (SOW).
August 2014                                                                         45

-------
Inorganic Data Review                                                                    ICP-MS
                               This page is intentionally left blank.
August 2014                                                                                  46

-------
Inorganic Data Review                                                                    ICP-MS

                            I.      Preservation and Holding Times
A.  Review Items
    Form 1-IN, Form 12-IN, Traffic Report/Chain of Custody (TR/COC) documentation, Form DC-1,
    raw data, and the Sample Delivery Group (SDG) Narrative checking for: pH; shipping container
    temperature; holding time; and other sample conditions.

B.  Objective
    The objective is to determine the validity of the analytical results based on sample conditions and the
    technical holding time of the sample.

C.  Criteria
    1.   The technical holding time is determined from the date of collection to the date of analysis.
    2.   The technical holding time criteria for aqueous/water samples is 180 days, preserved (with nitric
        acid) to pH < 2. The addition of nitric acid to adjust the pH is only required for aqueous/water
        samples.
    3.   The technical holding time criteria for soil/sediment samples is  180 days, based on the technical
        holding time criteria for aqueous/water samples.

    4.   Soil/sediment samples shall be maintained at < 6°C (but not frozen) from the time of collection
        until receipt at the laboratory. All aqueous/water and soil/sediment samples must be stored at
        < 6°C (but not frozen) from the time of sample receipt until digestion.
D.  Evaluation
    Establish technical holding times by comparing the sampling date(s) on the TR/COC documentation
    with the  dates of analysis on Form 12-IN and the raw data; also consider using information in the
    Complete SDG File (CSF), as it may be helpful in the assessment.  Verify that the analysis dates on
    the Form 12-IN and the raw data are identical. Review the SDG Narrative and raw data preparation
    logs to determine if samples were properly preserved. If there is an indication of problems with the
    samples, the  sample integrity may be compromised. Use professional judgment to evaluate the effect
    of the problem on the sample results.
E.  Action
    NOTE:    Apply the action to each field sample for which the preservation or holding time criteria
              was not met.
    1.   If the pH of aqueous/water metals samples is > 2 at the time of sample receipt, determine if the
        laboratory adjusted the pH to < 2 at the time of sample receipt.  If not, use professional judgment
        to qualify the samples based on the pH of the sample and the chemistry of the metal(s) of interest.
        Detects should be qualified as estimated low (J-) and non-detects as unusable (R).
    2.   If soil/sediment samples are not maintained at < 6°C (but not frozen) from the time of collection
        until receipt at the laboratory, detects should be qualified as estimated low (J-) and non-detects as
        unusable (R).
    3.   If technical holding times are exceeded, use professional judgment to determine the  reliability of
        the data based on the magnitude of the additional time compared to the technical requirement and
        whether the  samples were properly preserved. The expected bias would be low.  Detects should
        be qualified as estimated low (J-)  and non-detects as unusable (R).
    4.   Due  to limited information concerning holding times for soil/sediment samples, use  discretion
        when deciding whether to apply the aqueous/water holding time criteria to soil/sediment samples.
        If they are applied, annotate this in the Data Review Narrative.
August 2014                                                                                  47

-------
Inorganic Data Review
ICP-MS
    5.  If samples are received with shipping container temperatures > 10°C, use professional judgment
       to determine the reliability of the data, or qualify detects as estimated (J) and non-detects as
       estimated (UJ).
    6.  When the holding times are exceeded, annotate any possible consequences for the analytical
       results in the Data Review Narrative, and note it for Regional Laboratory Contracting Officer
       Representative (COR) action.

             Table 11. Preservation and Holding Time Actions for ICP-MS Analysis
Criteria
Aqueous/water samples received with pH > 2 and pH not
adjusted
Soil/sediment samples not maintained at < 6°C (but not
frozen) from time of collection until receipt at laboratory
Technical Holding Time:
Aqueous/water samples > 180 days
Technical Holding Time:
Soil/sediment samples > 180 days
Samples received > 10°C*
Action
Detect
Use professional
judgment
J-
Use professional
judgment
J-
J-
J-
Use professional
judgment
J
Non-detect
Use professional
judgment
R
Use professional
judgment
R
R
R
Use professional
judgment
UJ
       For samples received with shipping container temperatures > 10°C, Regional policy or project
       Data Quality Objectives (DQO) may allow the use of higher temperature criteria before assessing
       any actions for the affected samples.
August 2014
     48

-------
Inorganic Data Review                                                                    ICP-MS

                                     II.      Tune Analysis
A.  Review Items
    Form 13-IN, instrument printouts, and raw data.

B.  Objective
    The ICP-MS tune serves as an initial demonstration of instrument stability and precision.

C.  Criteria
    1.  Prior to calibration, the laboratory shall analyze or scan the ICP-MS tuning solution, containing
        100 (ig/L of Beryllium (Be), Magnesium (Mg), Cobalt (Co), Indium (In), and Lead (Pb), at least
       5x consecutively. The solution shall contain all required isotopes of these elements.  The
       laboratory shall make  any adjustments necessary to bring peak width within the instrument
       manufacturer's specifications and adjust the resolution of the mass calibration to within 0.1 (i
       over the range of 6-210 (i.
    2.  The Percent Relative Standard Deviation (%RSD) of the absolute signals for all analytes in the
       tuning solution must be < 5%.

D.  Evaluation
    1.  Verify, using the raw data and Form 13-IN, that the appropriate number of analyses or scans of
       the ICP-MS tuning solution were performed, and that the appropriate analytes were present in the
       solution.
    2.  Verify, using the raw data and Form 13-IN, that the resolution of the mass calibration falls within
       the limits for each isotope of each analyte.
    3.  Verify, using the raw data and Form 13-IN, that the %RSD is < 5% for each isotope of each
       analyte.
    4.  Verify, using the raw data, that the reported average mass and %RSD on Form 13-IN  was
       accurately calculated.  Recalculate one or more of the average masses and %RSDs for an isotope
       using the following equations:
                                                  Ix
                                           Mean = —
                                                   n
        Where,
               =   Mass from analysis
               =   Number of analyses
                                                              n.
                          Percent Relative Standard Deviation =
                                                                 x
        Where,
        x       =  Mean
        on_i     =  Standard Deviation

E.  Action
    NOTE:     For ICP-MS tunes that do not meet the technical criteria, apply the action to all samples
               reported from the analytical sequence.
    1.   If the ICP-MS instrument was not tuned prior to calibration, all sample data should be qualified
        as unusable (R).
August 2014                                                                                  49

-------
Inorganic Data Review
ICP-MS
    2.  If the tuning solution was not analyzed or scanned at least 5x consecutively, or the tuning solution
       does not contain the required analytes spanning the analytical range, use professional judgment to
       determine if the associated sample data should be qualified; obtain additional information from
       the laboratory, if necessary.  Record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action.
    3.  If the resolution of the mass calibration is not within 0.1 u for any isotope in the tuning solution,
       qualify the associated analytes that are detects as estimated (J) and non-detects as estimated (UJ).
       Record the situation in the Data Review Narrative,  and note it for Regional Laboratory COR
       action.

    4.  If the %RSD > 5% for any isotope in the tuning solution, qualify detects as estimated (J) and non-
       detects as estimated (UJ). Record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action.

                      Table 12. ICP-MS Tune Actions for ICP-MS Analysis
Criteria
Tune not performed
Tune not performed properly (Section II.E.2)
Resolution of mass calibration not within 0.1 u
%RSD > 5%
Action
Detect
R
Use professional
judgment
J
J
Non- detect
R
Use professional
judgment
UJ
UJ
August 2014
     50

-------
Inorganic Data Review                                                                     ICP-MS

                                       III.    Calibration
A.  Review Items
    Form 2-IN, Form 12-IN, Form 15-IN, Form 16-IN, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on initial calibration and
    calibration verification.

C.  Criteria
    1.  Initial Calibration
       The instruments shall be successfully calibrated each time the instrument is set up and after
       Continuing Calibration Verification (CCV) failure.  The calibration date and time shall be
       included in the raw data.
       a.  A blank and at least five calibration standards shall be used to establish each calibration
           curve. At least one standard shall be at or below the Contract Required Quantitation Limit
           (CRQL), but above the Method Detection Limit (MDL). All measurements shall be within
           the instrument working range. A minimum of three replicate scans are required for
           standardization for all Quality Control (QC) and sample analyses. The average result of all
           the multiple scans for the standardization, QC, and sample analyses shall be used. The
           calibration curve shall be fitted using  linear regression or weighted linear regression. The
           curve may be forced through zero. The curve must have a correlation coefficient of > 0.995.
           The calculated percent differences (%Ds) for all of the non-zero standards must be within
           ±30% of the true value of the standard. The y-intercept of the curve must be less than the
           CRQL.
    2.  Initial and Continuing Calibration Verification
       The acceptance criteria for the Initial Calibration Verification (ICV) and CCV standards are
       presented in Table 13:

         Table 13. Acceptance Criteria for ICV and CCV Standards for ICP-MS Analysis
Analytical Method
ICP-MS
Inorganic Analytes
Metals
ICV/CCV Low Limit
(% of True Value)
90
ICV/CCV High Limit
(% of True Value)
110
       a.  Initial Calibration Verification

           1)  Immediately after each system has been calibrated, the accuracy of the initial calibration
               must be verified and documented for each target analyte by the analysis of an ICV
               solution(s). If the ICV Percent Recovery (%R) falls outside of the control limits, the
               analysis should be terminated, the problem corrected, the instrument recalibrated, and all
               affected samples reanalyzed.
           2)  Only if the ICV is not available from the United States Environmental Protection Agency
               (EPA), analyses shall be conducted using a certified solution of the analytes from an
               independent commercial standard source, at a concentration level other than that used for
               instrument calibration, but within the calibrated range.
           3)  The ICV solution shall be analyzed at each analytical mass used for analysis.
August 2014                                                                                   51

-------
Inorganic Data Review                                                                    ICP-MS

       b.  Continuing Calibration Verification

           1)  To ensure accuracy during the course of each analytical sequence, the CCV shall be
               analyzed and reported for each mass used for the analysis of each analyte.

           2)  The CCV standard shall be analyzed at a frequency of every two hours during an
               analytical sequence. The CCV standard shall also be analyzed at the beginning of the
               analytical sequence, and again after the last analytical sample.
           3)  The analyte concentration(s) in the CCV standard(s) shall be different than the
               concentration used for the ICV, and at a concentration equivalent to the mid-level of their
               respective calibration curves.
           4)  The same CCV standard solution shall be used throughout the analysis for an SDG.

           5)  The CCV shall be analyzed in the same fashion as an actual sample.  If the %R of the
               CCV was outside of the control limits, the analysis should be terminated, the problem
               corrected, the instrument recalibrated, and all analytical samples analyzed since the last
               compliant CCV reanalyzed.
D.  Evaluation
    1.  Verify that the instrument was calibrated each time the instrument was set up, utilizing a blank
       and at least five calibration standards, one of which was at or below the CRQL, but above the
       MDL.
    2.  Confirm that the measurements were within the working calibration range, and were the average
       result of at least three replicate exposures.
    3.  Verify that the ICV and CCV standards were analyzed for each analyte at the specified frequency
       and at the appropriate concentration. Verify that acceptable %R results were obtained.
    4.  Recalculate one or more of the ICV and CCV %Rs using the following equation and verify that
       the recalculated value agrees with the laboratory-reported values on Form 2-IN.
                                         Found (value)
                                    %R=-—TT^XIO°
                                          I rue (value)

       Where,
       „    j /  i  x      Concentration (in ug/L) of each analyte measured in the analysis of the ICV or
       Found (value)   =  „„,,  i *•
              v    '      CCV solution
       True (value)    =  Concentration (in (ig/L) of each analyte in the ICV or CCV source
E.  Action
    NOTE:     For initial calibrations or ICV standards that do not meet the technical criteria,  apply the
               action to all associated samples reported from the analytical sequence.
               For CCV standards that do not meet the technical criteria, apply the action to all samples
               analyzed between a previous technically acceptable analysis of the QC sample and a
               subsequent technically acceptable analysis of the QC sample in the analytical sequence.
    1.  If the instrument was not calibrated each time the  instrument was set up, qualify detects and non-
       detects as unusable (R).  If the instrument was not calibrated with at least the minimum number of
       standards, or if the calibration curve does not include standards at required concentrations (e.g., a
       blank and at least one at or below CRQL but above MDL), use professional judgment to qualify
       detects as estimated (J) or unusable (R), and non-detects as estimated (UJ) or unusable  (R).
    2.  If the correlation coefficient is < 0.995, the %Ds are outside the ±30% limit, or the y-intercept >
       CRQL, qualify detects as estimated (J), and non-detects as estimated (UJ).
August 2014                                                                                  52

-------
Inorganic Data Review
ICP-MS
    3.  If the ICV or CCV %R falls outside the acceptance windows, use professional judgment to
       qualify all associated data. If possible, indicate the bias in the review. The following guidelines
       are recommended:
       a.  If the ICV or CCV %R is < 75%, use professional judgment to qualify detects as estimated
           low (J-) or unusable (R), and non-detects as unusable (R).
       b.  If the ICV or CCV %R falls within the range of 75-89%, qualify detects as estimated low (J-)
           and non-detects as estimated (UJ).
       c.  If the ICV or CCV %R falls within the range of 90-110%, detects and non-detects should not
           be qualified.
       d.  If the ICV or CCV %R falls within the range of 111-125%, qualify detects as estimated high
           (J+). Non-detects should not be qualified.
       e.  If the ICV or CCV %R is > 125%, use professional judgment to qualify detects as estimated
           high (J+) or unusable (R).  Non-detects should not be qualified.
    4.  If the laboratory failed to provide adequate calibration information, notify the Regional
       Laboratory COR. The Regional Laboratory COR may contact the laboratory and request the
       necessary information. If the information is unavailable, use professional judgment to assess the
       data.
    5.  Annotate the potential effects on the reported data due to exceeding the calibration criteria in the
       Data Review Narrative.
    6.  If calibration criteria are grossly exceeded, note this for Regional Laboratory COR action.
    NOTE:    For truly critical samples, a further in-depth evaluation of the calibration curve may be
               warranted to determine if additional qualification is necessary.

                       Table 14. Calibration Actions for ICP-MS Analysis
Criteria
Calibration not performed
Calibration incomplete
Correlation coefficient < 0.995; %D
outside ±30%; y-intercept > CRQL
ICV/CCV%R<75%
ICV/CCV %R 75-89%
ICV/CCV%R90-110%
ICV/CCV %R 1 1 1-125%
ICV/CCV %R> 125%
Action
Detect
R
Use professional judgment
JorR
J
Use professional judgment
J-orR
J
No qualification
J+
Use professional judgment
J+orR
Non-detect
R
Use professional judgment
UJorR
UJ
R
UJ
No qualification
No qualification
No qualification
August 2014
     53

-------
Inorganic Data Review                                                                    ICP-MS

                                         IV.    Blanks
A.  Review Items
    Form 1-IN, Form 3-IN, Form 12-IN, preparation logs, calibration standard logs, instrument logs, and
    raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the blank responses by
    determining the existence and magnitude of contamination resulting from laboratory (or field)
    activities or baseline drift during analysis.

C.  Criteria
    1.  No contaminants should be found in the blank(s).
    2.  The Initial Calibration Blank (ICB) shall be analyzed at each mass used for analysis after the
       analytical standards, but not before analysis of the ICV during the initial calibration of the
       instrument (see Section III.C.I).
    3.  A Continuing Calibration Blank (CCB) shall be analyzed at each mass used for the analysis,
       immediately after every CCV.  The CCB shall be analyzed at a frequency of every two hours
       during the analytical  sequence. The  CCB shall be analyzed at the beginning of the analytical
       sequence, and again after the last CCV that was analyzed after the last analytical sample of the
       analytical sequence.  The CCB result (absolute value) shall not exceed the CRQL of each analyte
       for which analysis is  performed.
    4.  At least one Preparation Blank shall  be prepared and analyzed for each matrix, with every SDG,
       or with each batch of samples digested, whichever is more frequent.  The Preparation Blank
       consists of reagent water processed through the appropriate  sample preparation and analysis
       procedure.
    5.  If the concentration of any analyte in the Preparation Blank  is  > CRQL, the lowest concentration
       of that analyte in the  associated samples must be > lOx the Preparation Blank concentration.
       Otherwise, all associated samples with the analyte's concentration <  lOxthe Preparation Blank
       concentration, and >  CRQL, should be redigested and reanalyzed for that analyte.  The laboratory
       is not to correct the sample concentration for the blank value.
    6.  If the concentration of any analyte in the Preparation Blank  is  < (-CRQL), all  associated samples
       with the analyte's concentration < lOx the CRQL should be redigested and reanalyzed.
D.  Evaluation
    1.  Verify that an ICB was analyzed after the calibration, the CCB was analyzed at the specified
       frequency and sequence during the analytical sequence, and Preparation Blanks were  prepared
       and analyzed as appropriate for the SDG (e.g., total number of samples, various types of matrices
       present, number of digestion batches, etc.).
    2.  Review the results reported on Form 3-IN, as well as the raw data for all blanks, and verify that
       the results were accurately reported.
    3.  Evaluate all of the associated blanks for the presence of target analytes. Verify that if the
       concentration of any target analyte was > CRQL in a Preparation Blank, all associated samples
       with analyte concentration > CRQL but < lOx the Preparation Blank concentration were
       redigested and reanalyzed for the analytes. Verify that if a concentration was < (-CRQL) in a
       Preparation Blank, all associated samples with the analyte's concentration < lOx CRQL were
       redigested and reanalyzed. Verify that if the absolute value of any target analytes was > CRQL in
       an ICB or a CCB, the analysis was terminated, the problem  corrected, the instrument  recalibrated,
       and the preceding 10 analytical samples or all analytical samples analyzed since the last
       compliant calibration blank reanalyzed.


August 2014                                                                                   54

-------
Inorganic Data Review                                                                   ICP-MS

E.  Action
    NOTES:   For ICBs that do not meet the technical criteria, apply the action to all associated samples
               reported from the analytical sequence.
               For CCBs that do not meet the technical criteria, apply the action to all associated
               samples analyzed between a previous technically acceptable analysis  of the CCB and a
               subsequent technically acceptable analysis of the CCB in the analytical sequence.
               For Preparation Blanks that do not meet the technical criteria, apply the action to all
               associated samples prepared in the same preparation batch.
    1.  If the appropriate blanks were not analyzed with the correct frequency, use professional judgment
       to determine if the associated sample data should be qualified; obtain additional information from
       the laboratory, if necessary. Record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action.

    2.  Action regarding unsuitable blank results depends on the circumstances and origin of the blank.
       In instances where more than one blank is associated with a given sample, qualification should be
       based upon a comparison with the associated blank having the highest concentration of
       contaminant.
    3.  Some general "technical" review actions include:
       a.   For any blank (including Preparation Blanks) reported with detects < CRQLs, report detects <
           CRQLs at the CRQLs and qualify as non-detect (U).  For any blank (including Preparation
           Blanks) reported with detects < CRQLs, use professional judgment to qualify the sample
           results > CRQL. Non-detects should not be qualified.
       b.   For any blank (including Preparation Blanks) reported with a negative result < (-MDL) but >
           (-CRQL), carefully evaluate  and determine its effect on the sample data.  Use professional
           judgment to assess the data.
       c.   The blank analyses may not involve the same weights, volumes, or dilution factors as the
           associated samples. In particular, soil/sediment sample results reported on Form 1-IN will
           not be on the same basis (units, dilution)  as the calibration blank data reported on Form 3-IN.
           It may be easier to work with the raw data and/or convert the ICB or CCB results to the same
           units as the soil/sediment samples for comparison purposes.
    4.  Specific "method" actions include:
       a.   If an ICB or a CCB result is > CRQL, the analysis should be terminated.  If the analysis was
           not terminated and the associated samples were not reanalyzed, non-detects should not be
           qualified.  Report detects < CRQLs at CRQLs and qualify as non-detect (U).  Report sample
           results that are > CRQLs but < ICB/CCB Results at ICB/CCB Results and use professional
           judgment to qualify as non-detect (U) or unusable (R). Use professional judgment to qualify
           sample results > ICB/CCB Results. Record the situation in the Data Review Narrative, and
           note it for Regional Laboratory COR action.

       b.   If an ICB or a CCB result is < (-CRQL), the analysis should be terminated. If the analysis
           was not terminated and the associated samples were not reanalyzed, use professional
           judgment to qualify non-detects as estimated (UJ) or unusable (R). Use professional
           judgment to qualify detects < CRQL, or qualify as estimated low (J-).  Use professional
           judgment to qualify sample results that are > CRQLs as estimated low (J-).
       c.   If the concentration of any analyte in the  Preparation Blank is > CRQL, the lowest
           concentration of that analyte in the associated samples must be > lOx the  Preparation Blank
           concentration. All samples associated  with the Preparation Blank with concentrations < lOx
           the Preparation Blank concentration and > CRQL should have been redigested and
           reanalyzed.  If the associated samples were not redigested and reanalyzed, report the sample
           results at Preparation Blank Results; use professional judgment to qualify the results as

August 2014                                                                                 55

-------
Inorganic Data Review
ICP-MS
           estimated high (J+) or unusable (R). Report detects < CRQLs in the samples associated with
           the Preparation Blank at CRQLs and qualify as non-detect (U).  Non-detects and sample
           results that are > lOx the Preparation Blank Results should not be qualified. If the laboratory
           failed to redigest and reanalyze the samples associated with the Preparation Blank, record it
           in the Data Review Narrative, and note it for Regional Laboratory COR action.
       d.  For any Preparation Blank reported with a negative result < (-CRQL), use professional
           judgment to qualify detects < CRQLs or qualify as estimated low (J-). Qualify sample results
           that are > CRQLs as estimated low (J-), and qualify non-detects as estimated (UJ). Sample
           results that are > lOx CRQLs should not be qualified.

                         Table 15. Blank Actions for ICP-MS Analysis
Blank
Type
ICB/CCB
ICB/CCB
ICB/CCB
ICB/CCB
Preparation
Blank
Preparation
Blank
Blank Result
Detect < CRQL
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Detect < CRQL
< (-MDL) but
> (-CRQL)
Sample Result
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but < ICB/CCB
Result
> ICB/CCB Result
Non-detect
Detect < CRQL
>CRQL
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Action
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at ICB/CCB Result and
qualify as non-detect (U) or unusable
(R)
Use professional judgment
Use professional judgment to qualify
as estimated (UJ) or unusable (R)
Use professional judgment or (J-)
Use professional judgment to qualify
as estimated low (J-)
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
August 2014
     56

-------
Inorganic Data Review
ICP-MS
Blank
Type
Preparation
Blank
Preparation
Blank
Blank Result
>CRQL
< (-CRQL)
Sample Result
Non-detect
Detect < CRQL
> CRQL but
< 1 Ox the Preparation Blank
Result
> 1 Ox the Preparation Blank
Result
Non-detect
Detect < CRQL
< lOx CRQL
> lOx CRQL
Action
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at Preparation Blank Result
and use professional judgment to
qualify results as estimated high (J+)
or unusable (R)
No qualification
Qualify as estimated (UJ)
Use professional judgment or (J-)
Report results > CRQL as estimated
low (J-)
No qualification
August 2014
     57

-------
Inorganic Data Review                                                                     ICP-MS

                               V.      Interference Check Sample
A.  Review Items
    Form 4-IN, Form 12-IN, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the instrument's ability to
    overcome interferences typical of those found in samples.

C.  Criteria
    1.  The Interference Check Sample (ICS) consists of two solutions: Solution A and Solution AB.
       Solution A consists of the interferents, and Solution AB consists of the analytes mixed with the
       interferents. An ICS analysis consists of analyzing both solutions consecutively, starting with
       Solution A, for all masses used for each analyte or interferent reported by ICP-MS.
    2.  An ICS must be analyzed undiluted at the beginning of each analysis sequence. The ICS is not to
       be analyzed prior to the ICV, and shall be immediately followed by a CCV, followed by a CCB.
    3.  Results for the analysis of the ICS Solution A must fall within the control limits of ± 2x the
       CRQL or ± 20% of the true value (whichever is greater) for the analytes and interferents included
       in the solution.
    4.  Results for the analysis of the ICS Solution AB must fall within the control limits of ± 2x the
       CRQL or ± 20% of the true value (whichever is greater) for the analytes and interferents included
       in the solution.
    5.  If the value of an ICS result exceeds ±  2x the CRQL, or ± 20% of true value (whichever is
       greater) criteria, the analysis shall be terminated, the problem corrected, the instrument
       recalibrated, the new calibration then reverified, and all analytical samples analyzed since the last
       compliant ICS reanalyzed.
    6.  The ICS should be obtained from the EPA, if available, and analyzed according to the
       instructions supplied with the solutions. If the ICS is not available from the EPA, an independent
       ICS solution shall be prepared using certified standards with the interferent and analyte
       concentrations at the levels specified in the method.
D.  Evaluation
    1.  Verify, using Form 12-IN and the raw data, that the ICS was analyzed at the specified frequency
       and sequence during the analytical sequence.
    2.  Evaluate the ICS  raw data for results with an absolute value that is > MDL for those analytes that
       are not present in the ICS solution.
    3.  Recalculate, using the raw data and the following equation, one  or more of the analyte %Rs, and
       verify that the recalculated value agrees with the laboratory-reported values on Form 4-IN.
                                         Found (value)
                                   %R=	/ ,   /  x 100
                                          True (value)
       Where,
       „     , ,  ,   ^      Concentration (in  ug/L) of each analyte interferent measured in the analysis
       Found (value)    =    ,,Tr,c c  ,  ..   .    T/c c  ,  .    AV>
                          of ICS Solution A or ICS Solution AB
       „    ,  ,  ,     _  Concentration (in  (ig/L) of each analyte or interferent in ICS Solution A or
        1 rue (value)     -  ICS Solutlon AB
August 2014                                                                                   58

-------
Inorganic Data Review                                                                      ICP-MS

    4.  If the value of an ICS result exceeds ± 2x the CRQL, or ± 20% of true value (whichever is
       greater) criteria, and the laboratory failed to terminate the analysis and take the appropriate
       corrective action, note this for Regional Laboratory COR action and record the situation in the
       Data Review Narrative. Use professional judgment to assess the data.
E.  Action
    NOTE:    For an ICS that does not meet the technical criteria, apply the action to all samples
               reported from the analytical sequence.
    1.  If the ICS was not analyzed at the specified frequency, qualify detects and non-detects as
       unusable (R). If the  ICS was analyzed, but not in the proper sequence, use professional judgment
       to qualify detects and non-detects.
    2.  The raw data may not contain results for interferents. In this  case, use professional judgment to
       qualify the data. If the data contains results for interferents, apply the following actions to
       samples with concentrations of interferents that are within 10% of the levels of the levels of
       interferent in the ICS:
       a.  If the ICS Solution AB %R for an analyte or interferent is < 50%, qualify detects as estimated
           low (J-) and non-detects as unusable (R).
       b.  If the ICS %R for an analyte or interferent falls within the range of 50-79% [or the ICS found
           value is < (true value - 2x the CRQL), whichever is lower], qualify detects as estimated low
           (J-), and non-detects as estimated (UJ).
       c.  If the ICS %R for an analyte or interferent falls within the range of 80-120%, detects and
           non-detects  should not be qualified.
       d.  The ICS %R for an analyte or interferent is > 120% [or the ICS found value is > (true value +
           2x the  CRQL), whichever is greater], qualify detects as estimated high (J+). Non-detects
           should not be qualified.
       e.  If the ICS %R for an analyte or interferent is above 150%, use professional judgment to
           determine the qualifications of the associated sample data.
    3.  If sample results that are > MDLs are observed for analytes which are not present in the ICS
       solution, the possibility of false positives exists. An evaluation of the associated sample data for
       the affected analytes should be made. For samples with comparable or higher levels of
       interferents and with analyte concentrations that approximate those levels found in the ICS,
       qualify detects as estimated high (J+). Non-detects should not be qualified.
    4.  If negative sample results are observed for analytes that are not present in the  ICS  solution, and
       their absolute values are > MDLs, the possibility of false negatives in the samples  exists. An
       evaluation of the associated sample data for the affected analytes should be made.  For samples
       with levels of interferents that are comparable to or higher than the levels found in the ICS,
       qualify detects < 1 Ox the  absolute value of the negative result as estimated low (J-), and qualify
       non-detects as estimated (UJ).
    NOTE:    The same result units should be used when comparing analyte results  in samples to those
               in the ICS.  Unit conversion may be necessary when  soil/sediment samples are  evaluated.
    5.  If the raw data does not contain results for the interferents, annotate this in the Data Review
       Narrative.
    6.  Actions regarding the interpretation and/or the  subsequent qualification of Inductively Coupled
       Plasma (ICP) data due to the ICS analytical results can be extremely complex. Use professional
       judgment to determine the need for the associated sample data to be qualified; obtain additional
       information from the laboratory, if necessary. Record all interpretive situations in the Data
       Review Narrative.
August 2014                                                                                    59

-------
Inorganic Data Review
ICP-MS
    7.  If the ICS acceptance criteria are grossly exceeded, note the specifics for Regional Laboratory
       COR action.
                   Table 16. Interference Check Actions for ICP-MS Analysis
Criteria
ICS not analyzed
ICS not analyzed in proper sequence
ICSAB%R<50%
ICS %R 50-79% [or ICS found value is < (true
value - 2x CRQL), whichever is lower]
ICS %R 80-120%
ICS%R> 150%
ICS %R > 120% [or ICS true value is > (true
value + 2x CRQL), whichever is greater]
Sample results > MDLs, but not present in ICS
Negative sample results, but not present in ICS
Action
Detect
R
Use professional
judgment
J-
J-
No qualification
Use professional
judgment
J+
J+
J-
for results
< 10x( negative sample
result)
Non- detect
R
Use professional
judgment
R
UJ
No qualification
Use professional
judgment
No qualification
No qualification
UJ
August 2014
     60

-------
Inorganic Data Review                                                                    ICP-MS

                               VI.     Laboratory Control Sample
A.  Review Items
    Form 7-IN, preparation logs, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the recovery of the
    digested Laboratory Control Sample (LCS).

C.  Criteria
    1.  Aqueous/water and soil/sediment LCSs shall be analyzed for each analyte utilizing the same
       sample preparations, analytical methods, and Quality Assurance/Quality Control (QA/QC)
       procedures as employed for the samples.
       a.   One LCS shall be prepared and analyzed for every group of aqueous/water or soil/sediment
           samples in an SDG, or with each batch of samples digested, whichever is more frequent. The
           LCS shall be spiked such that the final digestate contains each analyte at 2x the CRQL for the
           associated matrix.
       b.   All LCS %Rs must fall within the control limits of 70-130%. If the %R for the LCS falls
           outside of the control limits, the analysis should be terminated, the problem corrected, and the
           samples prepared with that LCS redigested and reanalyzed.
D.  Evaluation
    1.  Verify, using Form 7-IN, preparation logs, and raw data, that the appropriate number of required
       LCSs were prepared and analyzed for the SDG.
    2.  Verify, using Form 7-IN, that all results for each analyte fall within the established control limits.
       a.   Check the raw data to verify that the %Rs on Form 7-IN were accurately transcribed.
           Recalculate one or more of the reported %Rs using the following equation:

                                         Found (value)
                                    %R=	 xlOO
                                          True (value)

        Where,
        „    j /  i  x       Concentration of each analyte (in ug/L or mg/kg) measured in the analysis
        Found (value)    =   ,,.,   T „„               J  \  rs      55;                   j
                           of the LCS
        True (value)     =  Concentration of each analyte (in (ig/L or mg/kg) in the LCS

    3.  Verify that the LCS was prepared at the same time as the associated samples using the same
       procedures.
E.  Action
    NOTE:    If the LCS criteria are not met, the laboratory performance and method accuracy are in
              question.  Use professional judgment to determine if the data should be qualified or
              rejected. The following guidance is suggested for qualifying sample data associated with
              an LCS that does not meet the required criteria.
              For an LCS that does not meet the technical criteria, apply the action to all samples in the
              same preparation batch.
    1.  If the required LCS was not analyzed at the specified frequency, use professional judgment to
       determine if the associated sample data should be qualified; obtain additional information from
       the laboratory, if necessary. If a laboratory fails to analyze a LCS with each SDG, or if a
       laboratory consistently fails to generate acceptable LCS recoveries, record the situation in the


August 2014                                                                                  61

-------
Inorganic Data Review
ICP-MS
       Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ).

    2.  LCS for all matrices:

       a.  If LCS %R is < 40%, qualify detects as estimated low (J-) and non-detects as unusable (R).

       b.  If the LCS %R falls within the range of 40-69%, qualify detects as estimated low (J-) and
           non-detects as estimated (UJ).
       c.  If the LCS %R falls within the range of 70-130%, detects and non-detects should not be
           qualified.

       d.  If the LCS %R is > 130%, qualify detects as estimated high (J+).  Non-detects should not be
           qualified.

       e.  If the LCS %R is > 150%, qualify detects as unusable (R).  Non-detects should not be
           qualified.
    3.  Annotate the potential effects on the data due to out-of-control LCS results in the Data Review
       Narrative.

                          Table 17. LCS Actions for ICP-MS Analysis
Criteria
Aqueous/Water and Soil/Sediment %R < 40%
Aqueous/Water and Soil/Sediment %R 40-69%
Aqueous/Water and Soil/Sediment %R 70-130%
Aqueous/Water and Soil/Sediment %R > 130%
Aqueous/Water and Soil/Sediment %R > 150%
Action
Detect
J-
J-
No qualification
J+
R
Non- detect
R
UJ
No qualification
No qualification
No qualification
August 2014
     62

-------
Inorganic Data Review                                                                   ICP-MS

                               VII.   Duplicate Sample Analysis
A.  Review Items
    Cover Page, Form 6-IN, instrument printouts, and raw data.

B.  Objective
    The objective of duplicate sample analysis is to demonstrate acceptable method precision by the
    laboratory at the time of analysis.

C.  Criteria
    1.   Samples identified as field blanks or Performance Evaluation (PE) samples cannot be used for
        duplicate sample analysis.
    2.   At least one duplicate sample shall be prepared and analyzed from each group of samples of a
        similar matrix type (e.g., water or soil) or for each SDG. Duplicates cannot be averaged for
        reporting on Form 1-IN. Additional duplicate sample analyses may be required by EPA Regional
        request.  Alternately, the Region may require that a specific sample be used for the duplicate
        sample analysis.
    3.   A control limit of 20% for the Relative Percent Difference (RPD) shall be used for original and
        duplicate sample values > 5x the CRQL.
    4.   A control limit of the CRQL shall be used if either the sample or duplicate value is < 5x the
        CRQL.  The absolute value of the control limit  (CRQL)  shall be entered in the "Control Limit"
        column on Form 6-IN.  If both samples are non-detects, the RPD is not calculated for Form 6-IN.
    NOTE:     The above control limits are method requirements for duplicate samples, regardless of
               the sample matrix type. However, it should be noted that laboratory variability arising
               from the sub-sampling of non-homogenous soil  samples is a common occurrence.
               Therefore, for technical review purposes only, Regional policy or project DQOs may
               allow the use of less restrictive criteria  (e.g.,  35% RPD, 2x the CRQL) to be assessed
               against duplicate soil samples.
D.  Evaluation
    1.   Verify, from the Cover Page and the raw data, that the appropriate number of required duplicate
        samples were prepared and analyzed for the SDG.
    2.   Verify, using Form 6-IN and the raw data, that  all duplicate results for each analyte fall within the
        established control limits.
    3.   Verify that a field blank or PE sample was not used for duplicate analysis.
    4.   Check the raw data and recalculate one or more of the RPD values using the following equation
        to verify that the results were correctly reported on Form 6-IN:
                                    RPD
                                              S-D
                                            (S + D)/2
        Where,
        RPD   =   Relative Percent Difference
        S      =   Sample Result (original)
        D      =   Duplicate Result
100
August 2014                                                                                 63

-------
Inorganic Data Review                                                                    ICP-MS

E.  Action
    NOTE:     For a duplicate sample analysis that does not meet the technical criteria, apply the action
               to all samples of the same matrix if the samples are considered sufficiently similar.
               Exercise professional judgment in determining sample similarity when making use of all
               available data, including: site and sampling documentation (e.g., location and type of
               sample, descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity,
               chlorine); and laboratory data for other parameters [e.g., Total Suspended Solids (TSS),
               Total Dissolved Solids (TDS), Total Organic Carbon (TOC), alkalinity or buffering
               capacity, reactive sulfide, anions].  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG.
               Possible determinations are:  1) only some of the samples in the SDG are similar to the
               duplicate sample, and that only these samples should be qualified; or, 2) no samples are
               sufficiently similar to the sample used for the duplicate, and thus only the field sample
               used to prepare the duplicate sample should be qualified.
    1.  If the appropriate number of duplicate samples were not analyzed for each matrix using the
       correct frequency, use professional judgment to  determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary.  Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If both original sample and duplicate sample results are  > 5x the CRQL and the RPD is > 20%,
       qualify detects as estimated (J), and qualify non-detects  as estimated (UJ).
    3.  If RPD > 100%,  use professional judgment to determine if the associated sample data should be
       qualified.
    4.  If both original sample and duplicate sample results are  > 5x the CRQL and the RPD is < 20%,
       detects and non-detects should not be qualified.
    5.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate > CRQL, qualify detects as estimated (J) and
       non-detects as estimated (UJ).
    6.  If the original sample or duplicate sample result is < Sxthe CRQL (including non-detects) and
       the absolute difference between sample and duplicate <  CRQL, detects and non-detects should
       not be qualified.
    7.  If a field blank or PE sample was used for the duplicate  sample  analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be  carefully checked. Use
       professional judgment when evaluating the data.
    8.  Annotate the potential effects  on the  data due to out-of-control duplicate sample results in the
       Data Review Narrative.
August 2014                                                                                  64

-------
Inorganic Data Review
ICP-MS
                    Table 18. Duplicate Sample Actions for ICP-MS Analysis
Criteria
Both original sample and duplicate sample results are > 5x
the CRQL and RPD > 20%*
Both original sample and duplicate sample results are > 5x
the CRQL and RPD is < 20%
RPD > 100%
Original sample or duplicate sample result < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate > CRQL*
Original sample or duplicate sample result < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate < CRQL
Action
Detect
J
No qualification
Use professional
judgment
J
No qualification
Non-detect
UJ
No qualification
Use professional
judgment
UJ
No qualification
    *   The above control limits are method requirements for duplicate samples, regardless of the
       sample matrix type. However, it should be noted that laboratory variability arising from the sub-
       sampling of non-homogenous soil samples is a common occurrence.  Therefore, for technical
       review purposes only, Regional policy or project DQOs may allow the use of less restrictive
       criteria (e.g., 35% RPD, 2xthe CRQL) to be assessed against duplicate soil samples.
August 2014
     65

-------
Inorganic Data Review                                                                    ICP-MS

                                 VIII.  Spike Sample Analysis
A.  Review Items
    Cover Page, Form 5A-IN, Form 5B-IN, instrument printouts, and raw data.

B.  Objective
    The objective of the spiked sample analysis is to evaluate the effect of each sample matrix on the
    sample preparation procedures and the measurement methodology.

C.  Criteria
    1.   Samples identified as field blanks or PE samples cannot be used for spiked sample analysis.
    2.   At least one spiked sample (pre-digestion) shall be prepared and analyzed from each group of
        samples with a similar matrix type (e.g., water or soil), or for each SDG.

    3.   When the Matrix Spike recovery falls outside of the control limits and the sample result is < 4x
        the spike added, a post-digestion spike shall be performed for those analytes that do not meet the
        specified criteria.  An aliquot of the remaining unspiked sample shall be spiked at 2x the
        indigenous level or 2x the CRQL, whichever is greater.

    4.   The spike %R shall be within the established acceptance limits. However, spike recovery limits
        do not apply when the sample concentration is > 4x the spike added. In such an event, the data
        shall be reported unflagged, even if the %R does not meet the acceptance criteria.
    5.   If the spiked sample analysis was performed on the same sample that was chosen for the duplicate
        sample analysis, spike calculations shall be performed using the results of the sample designated
        as the "original sample." The average of the duplicate  results cannot be used for the purpose of
        determining %R.

    NOTE:     The final spike  concentrations required for the  various target analytes are presented in the
               methods described in the SOW.
D.  Evaluation
    1.   Verify, using the Cover Page, Form 5 A-IN, and raw data, that the appropriate number of required
        spiked samples was prepared and analyzed for the SDG.
    2.   Verify that a field blank or PE sample was not used for the spiked sample analysis.
    3.   Verify, using Form 5A-IN and the raw data, that all pre-digestion spiked sample  results for each
        required analyte fall within the established control limits.  If not, verify that a post-digestion spike
        was prepared and analyzed.
    4.   Recalculate using the raw data, one or more of the %R using the following equation, and verify
        that the recalculated value agrees with the laboratory-reported values on Forms 5A-IN & 5B-IN:
                                                SSR-SR
                                  %Recovery=  ——— *100
        Where,
        SSR   =   Spiked Sample Result
        SR     =   Sample Result
        SA     =   Spike Added
    NOTE:     When the sample result is < MDL or reported as a non-detect, use SR = 0 only for the
               purpose of calculating the %R. The actual spiked sample results, sample results, and %R
               (positive or negative) shall still be reported on Forms 5A-IN & 5B-IN.
August 2014                                                                                  66

-------
Inorganic Data Review                                                                     ICP-MS

E.  Action
    NOTE:     For a Matrix Spike that does not meet the technical criteria, apply the action to all
               samples of the same matrix, if the samples are considered sufficiently similar.  Exercise
               professional judgment in determining sample similarity when making use of all available
               data, including: site and sampling documentation (e.g., location and type of sample,
               descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity, chlorine);
               and laboratory data for other parameters (e.g., TSS, TDS, TOC, alkalinity or buffering
               capacity, reactive sulfide, anions).  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG.
               Possible determinations are: 1) only some of the samples in the SDG are similar to the
               Matrix Spike sample, and that only these samples should be qualified; or, 2) no samples
               are sufficiently similar to the sample used for the Matrix Spike, and thus only the field
               sample used to prepare the Matrix  Spike sample should be qualified.
    1.  If the appropriate number of Matrix Spike samples was not analyzed for each matrix using the
       correct frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary.  Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If a field blank or PE sample was used for the spiked sample analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be carefully checked.  Use
       professional judgment when evaluating the data.  Detects should be qualified as estimated (J) and
       non-detects as estimated (UJ).
    3.  If the Matrix Spike recovery does not meet the evaluation criteria and a required post-digestion
       spike was not performed, note this for Regional Laboratory COR action.
    4.  If the Matrix Spike %R is < 30%, verify that a post-digestion spike was analyzed. If the post-
       digestion spike %R is < 75% or the analysis was not performed, qualify detects as estimated low
       (J-) and non-detects as unusable (R). If the post-digestion  spike %R is > 75%, qualify detects as
       estimated (J) and non-detects as estimated (UJ).
    5.  If the Matrix Spike %R falls within the range of 30-74%, verify that a post-digestion spike was
       analyzed (if required when sample concentration is < 4x spike added).  If the post-digestion spike
       %R is < 75% or the analysis was not performed, qualify detects as estimated low (J-) and non-
       detects as estimated (UJ).  If the post-digestion spike %R for is > 75%, qualify detects as
       estimated (J) and non-detects as estimated (UJ).
    6.  If the Matrix Spike %R falls within the range of 75-125%, no post-digestion spike is required.
       Detects and non-detects should not be qualified.
    7.  If the Matrix Spike %R is > 125%, verify that a post-digestion spike was analyzed (if required
       when sample concentration is < 4x spike added).  If the post-digestion spike %R is also > 125%
       or the analysis was not performed, qualify detects as estimated high (J+); non-detects should not
       be qualified. If the post-digestion spike %R is < 125%,  qualify detects as  estimated (J); non-
       detects should not be qualified.
    8.  Annotate the potential effects on the data due to out-of-control spiked sample results in the Data
       Review Narrative.
August 2014                                                                                   67

-------
Inorganic Data Review
ICP-MS
                      Table 19. Spike Sample Actions for ICP-MS Analysis
Criteria
Matrix Spike %R < 30%
Post-digestion spike %R < 75%
Matrix Spike %R < 30%
Post-digestion spike %R> 75%
Matrix Spike %R 30-74%
Post-digestion spike %R < 75%
Matrix Spike %R 30-74%
Post-digestion spike %R> 75%
Matrix Spike %R> 125%
Post-digestion spike %R> 125%
Matrix Spike %R> 125%
Post-digestion spike %R< 125%
Matrix Spike %R < 30%
No post-digestion spike performed
Matrix Spike %R 30-74%
No post-digestion spike performed
Matrix Spike %R 75-125%
No post-digestion spike is required
Matrix Spike %R> 125%
No post-digestion spike performed
Action
Detect
J-
J
J-
J
J+
J
J-
J-
No qualification
J+
Non- detect
R
UJ
UJ
UJ
No qualification
No qualification
R
UJ
No qualification
No qualification
    NOTE:    The above control limits are method requirements for spike samples, regardless of the
              sample matrix type. However, it should be noted that laboratory variability arising from
              the sub-sampling of non-homogenous soil samples is a common occurrence. Therefore,
              for technical review purposes only, Regional policy or project DQOs may allow the use
              of less restrictive criteria (e.g., 10 %R and 150 %R for the lower and upper limits) to be
              assessed against spike and post-digestion spike soil samples.
August 2014
     68

-------
Inorganic Data Review                                                                     ICP-MS

                                     IX.     Serial Dilution
A.  Review Items
    Form 1-IN, Form 8 -IN, instrument printouts, and raw data.

B.  Objective
    The objective of the serial dilution analysis is to determine whether or not significant physical or
    chemical interferences exist due to sample matrix.

C.  Criteria
    1 .  An ICP Serial Dilution analysis shall be performed on a sample from each group of samples with
       a similar matrix type (e.g., water or soil) or for each SDG, whichever is more frequent.
    2.  Samples identified as field blanks or PE samples cannot be used for the ICP Serial Dilution
       analysis.
    3.  If the analyte concentration is sufficiently high (concentration in the original sample is > 5 Ox the
       MDL), the %D between the original determination and the serial dilution analysis (a five-fold
       dilution) after correction for dilution (concentration in the serial dilution sample is > CRQL) shall
       be < 10%.

    NOTE:    The above criteria are method requirements for serial dilution samples, regardless of the
               sample matrix type. However, for technical review purposes only, Regional policy or
               project DQOs may allow the use of less restrictive criteria (e.g., %D > 15%) to be
               assessed against serial dilution soil samples.
D.  Evaluation
    1 .  Verify that a field blank or PE sample was not used for the serial dilution analysis.
    2.  Check the raw data and recalculate the %D using the following equation. Verify that the serial
       dilution analysis results and the calculated %D results agree with the values reported by the
       laboratory on Form 8 -IN:
                                                    -
                                    %Difference =
        Where,
        I       =   Initial Sample Result
        S      =   Serial Dilution Result

    3.   Check the raw data for any evidence of positive or negative interference (results from the diluted
        sample which are significantly different than the original sample), possibly due to high levels of
        dissolved solids in the sample, ionization effects, etc.
E.  Action
    NOTE:     For a serial dilution that does not meet the technical criteria, apply the action to all
               samples of the same matrix if the samples are considered sufficiently similar.  Exercise
               professional judgment in determining  sample similarity when making use of all available
               data, including: site and sampling documentation (e.g., location and type of sample,
               descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity, chlorine);
               and laboratory data for other parameters (e.g., TSS, TDS, TOC, alkalinity or buffering
               capacity,  reactive sulfide, anions).  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG. Two
               determinations are: 1) only some of the samples in the SDG are similar to the serial
               dilution sample, and that only these samples should be qualified; or, 2) no samples are
August 2014                                                                                   69

-------
Inorganic Data Review
ICP-MS
               sufficiently similar to the sample used for serial dilution, and thus only the field sample
               used to prepare the serial dilution sample should be qualified.
    1.  If the appropriate number of serial dilution samples was not analyzed for each matrix, use
       professional judgment to determine if the associated sample data should be qualified; obtain
       additional information from the laboratory, if necessary.  If a field blank or PE sample was used
       for the serial dilution analysis, record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action. Detects should be qualified as estimated (J) and non-detects as
       estimated (UJ).
    2.  If the analyte concentration in the original sample is > 5 Ox the MDL, its concentration in the
       serial dilution sample is > CRQL, and the %D > 10%, qualify detects as estimated (J) and non-
       detects as estimated (UJ).
    3.  If the analyte concentration in the original sample is > 5 Ox the MDL, its concentration in the
       serial dilution sample is > CRQL, and the %D is < 10%,  detects and non-detects should not be
       qualified.
    4.  If the analyte concentration in the original sample is > 5 Ox the MDL, its concentration in the
       serial dilution sample is > CRQL, and %D is > 100%, use professional judgment to determine if
       the associated sample data should be qualified.
    5.  If the analyte concentration in the original sample is > 5x the CRQL and its concentration in the
       serial dilution sample is < CRQL, detects and non-detects should not be qualified.
    6.  If evidence of positive or negative interference is found, use professional judgment to qualify the
       associated sample data. Annotate the potential effects on the reported data in the Data Review
       Narrative.
                      Table 20.  Serial Dilution Actions for ICP-MS Analysis
Criteria
Sample concentration
> 5 Ox MDL, serial dilution sample
concentration > CRQL, and %D >
10%*
Sample concentration
> 5 Ox MDL, serial dilution sample
concentration > CRQL, and %D <
10%
Sample concentration
> 5 Ox MDL, serial dilution sample
concentration > CRQL, and %D >
100%
Sample concentration > 5x CRQL and
serial dilution sample concentration <
CRQL
Interferences present
Action
Detect
J
No qualification
Use professional
judgment
No qualification
Use professional
judgment
Non-detect
UJ
No qualification
Use professional
judgment
No qualification
Use professional
judgment
       The above criteria are method requirements for serial dilution samples, regardless of the sample
       matrix type. However, for technical review purposes only, Regional policy or project DQOs
       may allow the use of less restrictive criteria (e.g., %D > 15%) to be assessed against serial
       dilution soil samples.
August 2014
     70

-------
Inorganic Data Review                                                                      ICP-MS

                                    X.     Internal Standards
A.  Review Items
    Form 11-IN, Form 14-IN, instrument printouts, and raw data.

B.  Objective
    The objective of internal standard analysis is to determine the existence and magnitude of instrument
    drift and physical interferences.

C.  Criteria
    1.   All samples analyzed during an analytical sequence, with the exception of the tune, shall contain
        internal standards. A minimum of five internal standards from the following list shall be added to
        each sample: Li (the Li6 isotope); Sc; Y; Rh; In; Tb; Ho; Lu; and Bi. If the laboratory uses
        lithium as an internal standard, the laboratory shall use an Li6-enriched standard. The laboratory
        shall monitor the same internal standards throughout the entire analytical sequence and shall
        assign each analyte to at least one internal standard.
    2.   The intensity of the internal standard response in a sample is monitored and compared to the
        intensity of the response for that internal standard in the calibration blank. The Percent Relative
        Intensity (%RI) in the sample shall fall within 60-125% of the response in the calibration blank.
    3.   If the %RI of the response in the sample falls outside of these  limits, the laboratory shall
        reanalyze the original sample at a two-fold dilution with internal standard added.
D.  Evaluation
    1.   Verify, using Form 14-IN and the raw data, that a minimum of five internal standards from the
        specified list were used for the analysis; that the same internal standards were monitored for the
        entire analytical sequence; and that each analyte was associated to at least one internal standard.
    2.   Verify, using Form 14-IN and the raw data, that these internal standards were added to each
        sample in the analytical sequence, including calibrations, samples, and QC samples (except tune).
    3.   Verify, using Form 14-IN, that the %RI between an internal standard in a sample and the internal
        standard in the calibration blank was reported for each sample.
    4.   Verify, using Form 11-IN, Form 14-IN, and the raw data, that if the %RI for a sample was outside
        the limits (60-125%), the sample was reanalyzed of a two-fold dilution with internal standard
        added.
E.  Action
    NOTE:    Apply the action to the affected analytes for each sample that does not meet the internal
               standard criteria.
    1.   If no internal standards were analyzed with the analytical sequence, detects and non-detects
        should be qualified as unusable (R).  Record this issue in the Data Review Narrative, and note it
        for Regional Laboratory COR action.

    2.   If less than five of the required internal standards were analyzed with the analytical sequence, or
        (a) target analyte(s) is (are) not associated to an internal standard, the sample results, for the
        analyte(s) not associated to an internal standard should be qualified as unusable (R).  Record this
        issue in the Data Review Narrative, and note it for Regional Laboratory COR action.
    3.   If the %RI for the internal standards in a sample falls within the range of 60-125%, the sample
        results should not be qualified.
    4.   If the %RI for an internal standard in a sample is < 60% or > 125%, qualify the sample results of
        the analytes associated with the non-compliant internal standard(s) as follows:
August 2014                                                                                    71

-------
Inorganic Data Review
                                                                               ICP-MS
       b.
If the sample was reanalyzed at a two-fold dilution with internal standard %RIs within the
limits, report the result (for those analytes associated to the internal standard outside the
limits in the initial analysis) from the diluted analysis without qualification.  If any of the
%RIs of the diluted analysis  were < 60% or > 125%, report the results of the original
undiluted analysis and qualify the detects as estimated  (J) and non-detects as estimated (UJ)
for the associated analytes.
If the sample was not reanalyzed at a two-fold dilution, use professional judgment to
determine the reliability of the data. Detects should be qualified as estimated (J) or unusable
(R) and non-detects as estimated (UJ) or unusable (R) for the associated analytes.

         Table 21. Internal Standard Actions for ICP-MS Analysis
Criteria
No internal standards
< 5 of the required internal standards
Target analyte not associated with internal standard
%RI 60-125%
%RI < 60% or > 125% and original sample
reanalyzed at 2-fold dilution
Original sample not reanalyzed at 2-fold dilution
Action
Detect
R
R
R
No qualification
J
Use professional
judgment
JorR
Non-detect
R
R
R
No qualification
UJ
Use professional
judgment
UJorR
August 2014
                                                                                    72

-------
Inorganic Data Review                                                                  ICP-MS

                    XI.     Regional Quality Assurance and Quality Control
A.  Review Items
    Form 1 -IN, instrument printouts, and raw data.

B.  Objective
    The objective is to use results from the analysis of Regional QA/QC samples such as field blanks, PE
    samples, blind spikes, and blind blanks to determine the validity of the analytical results.

C.  Criteria
    Criteria are determined by each Region.
D.  Evaluation
    Evaluation procedures must follow the Region's Standard Operating Procedure (SOP) for data
    review. Each Region will handle the evaluation of PE samples on an individual basis. Compare
    results for PE samples to the acceptance criteria for the specific PE samples if possible.
    Calculate the RPD between field duplicates and provide this information in the Data Review
    Narrative.
E.  Action
    Any action must be  in accordance with Regional specifications and criteria for acceptable PE sample
    results. Note any unacceptable PE sample results for Regional Laboratory COR action.
August 2014                                                                                 73

-------
Inorganic Data Review                                                                    ICP-MS

                               XII.    Overall Assessment of Data
A.  Review Items
    Entire sample data package, data review results, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data (including any confirmation data).

B.  Objective
    The objective is to provide the overall assessment on data quality and usability.

C.  Criteria
    1.  Review all available materials to assess the overall quality of the data, keeping in mind the
       additive nature of analytical problems.
    2.  Reported analyte concentrations must be quantitated according to the appropriate analytical
       method, as listed in the method. All sample results must be within the linear calibration ranges
       per methods. Percent Solids (%Solids) must be properly used for all applicable matrix result
       calculations.
D.  Evaluation
    Examine the raw data to verify that the correct calculation of the sample results was reported by the
    laboratory. Digestion logs, instrument printouts, etc., should be compared to the reported sample
    results recorded on the appropriate Inorganic Summary Forms (Form 1-IN through Form 16-IN).
    1.  Evaluate any technical problems not previously addressed.
    2.  Examine the raw data for any anomalies (e.g., baseline shifts, negative response, mass dependent
       drift, omissions, illegibility, etc.).
    3.  Verify that appropriate methods and volumes were used in preparing the samples for analysis.  If
       reduced volumes were used, verify that the laboratory had received Regional Laboratory COR
       approval for the  use of the reduced volume.
    4.  Verify that there are no transcription or reduction errors (e.g., dilutions, %Solids, sample weights,
       etc.) on one or more samples. Recalculate %Solids for at least 10% of the samples and verify that
       the calculated %Solids agree with that reported by the library.
    5.  Verify that MDLs are properly reported and that they are not greater than the respective CRQLs.
    6.  Verify that results fall within the calibrated range(s) of the instrument(s) (Form 15-IN).
    7.  If appropriate information is available, assess the usability of the data to assist the data user in
       avoiding inappropriate use of the data.  Review all available information, including the Quality
       Assurance Project Plan (QAPP), focusing specifically on the acceptance or performance criteria,
       the SOPs, and communication with the user concerning the intended use and desired quality of
       these data.
E.  Action
    1.  Use professional judgment to determine if there is any need to qualify data which were not
       qualified based on the QC criteria previously discussed.
    2.  Use professional judgment to qualify detects and non-detects if the MDL exceeds CRQL.
    3.  If a sample is not diluted properly when sample results exceed the upper limit of the calibration
       range, qualify detects as estimated (J).
    4.  Write a brief Data Review Narrative to give the user an indication of the analytical limitations  of
       the data. Annotate any discrepancies between the data and the SDG Narrative for Regional
       Laboratory COR action. If sufficient information on the intended use and required quality of the
       data is available, include an assessment of the data usability within the given context.
August 2014                                                                                    74

-------
Inorganic Data Review                                                                    ICP-MS

    5.  If any discrepancies are found, notify the Regional Laboratory COR.  The Regional Laboratory
       COR may contact the laboratory to obtain additional information for a resolution. If a
       discrepancy remains unresolved, use professional judgment to determine if qualification of the
       data is warranted.
August 2014                                                                                   75

-------
Inorganic Data Review                                                                  ICP-MS

                                     XIII.   Calculations
Aqueous/Water Sample Concentration by ICP-MS:
    The concentrations determined in the digestate are to be reported in units of (ig/L:
    Where,
                                               .        Vf
                              Concentration (ug/L) = C x — x DF
    C      =  Instrument value in |o,g/L (the average of all replicate integrations)
    Vf     =  Final digestion volume (mL)
    V      =  Initial Aliquot Amount (mL)
    DF    =  Dilution Factor

Soil/Sediment Sample Concentration by ICP-MS:
    The concentrations determined in the digestate are to be reported on the basis of the dry weight of the
    sample, in units of mg/kg:
                                                            Vf
                     Concentration (mg/kg dry weight) = C x ——- x DF/1000
                                                          W x b
    Where,
    C      =  Instrument value in |o,g/L (the average of all replicate integrations)
    Vf     =  Final digestion volume (mL)
    W     =  Initial aliquot amount (g)
    S      =  %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    =  Dilution Factor

Adjusted MDL/Adjusted CRQL Calculation:
    To calculate the adjusted MDL or adjusted CRQL for aqueous/water samples, substitute the value of
    the MDL ((ig/L) or CRQL ((ig/L) into the "C" term in the equation above.

    Calculate the adjusted MDL or adjusted CRQL for soil/sediment samples as follows:
                                                         WM    Vf
                    Adjusted MDL or CRQL (mg/kg) = C x —^- x _i  x DF
                                                        W x S   V
    Where,
    C      =  MDL or CRQL (mg/kg)
    WM    =  Minimum method required aliquot amount (g) (1 .OOg or 0.50g)
    W     =  Initial aliquot amount (g)
    VM    =  Method required final sample digestion volume (mL) (100 mL)
    Vf     =  Final digestion volume  (mL)
    S      =  %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    =  Dilution Factor
August 2014                                                                                 76

-------
Inorganic Data Review                                                                  Mercury
                                 MERCURY DATA REVIEW
The inorganic data requirements for mercury to be reviewed during validation are listed below:
Example Analytical Sequence	79
I.    Preservation and Holding Times	81
II.   Calibration	83
III.  Blanks	86
IV.  Duplicate Sample Analysis	90
V.   Spike Sample Analysis	93
VI.  Regional Quality Assurance and Quality Control	95
VII.  Overall Assessment of Data	96
VIII. Calculations	98
August 2014                                                                                 77

-------
Inorganic Data Review                                                                    Mercury
                                This page is intentionally left blank.
August 2014                                                                                   78

-------
Inorganic Data Review                                                             Mercury

                              Example Analytical Sequence
This is an example of an analytical sequence:

S##
S##
s##
s##
s##
s##
ICV
ICB
ccv###
CCB###
samples
ccv###
CCB###
samples
ccv###
CCB###, etc.
*Suffix ## and ### are as specified in Exhibit B of the Statement of Work (SOW).
August 2014                                                                          79

-------
Inorganic Data Review                                                                    Mercury
                                This page is intentionally left blank.
August 2014                                                                                   80

-------
Inorganic Data Review                                                                   Mercury

                              I.   Preservation and Holding Times
A.  Review Items
    Form 1-IN, Form 12-IN, Traffic Report/Chain of Custody (TR/COC) documentation, Form DC-1,
    raw data, and the Sample Delivery Group (SDG) Narrative checking for: pH; shipping container
    temperature; holding time; and other sample conditions.

B.  Objective
    The objective is to determine the validity of the analytical results based on the sample conditions and
    the holding time of the sample.

C.  Criteria
    1.   The technical holding time is determined from the date of collection, or the date Toxicity
        Characteristic Leaching Procedure (TCLP) or Synthetic Precipitation Leaching Procedure (SPLP)
        extraction is complete, to the date of analysis.
    2.   The technical holding time criteria for aqueous/water samples and leachate samples from TCLP
        or SPLP is 28 days, preserved (with nitric acid) to pH < 2.
    3.   The technical holding time criteria for soil/sediment samples is 28 days, based on the technical
        holding time criteria for aqueous/water samples.
    4.   Soil/sediment samples shall be maintained at < 6°C (but not frozen) from the time of collection
        until receipt at the laboratory. All aqueous/water and soil/sediment samples must be stored at
        < 6°C (but not frozen) from the time of sample receipt until digestion. The TCLP and SPLP
        leachates must be stored at < 6°C (but not frozen) from the time of the leaching procedure
        completion until digestion.

    5.   Samples and standards shall be analyzed with 48 hours of preparation.
D.  Evaluation
    Establish technical holding times by comparing the sampling date(s) on the TR/COC documentation
    with the  dates of analysis on Form 12-IN and the raw data; also consider using information in the
    Complete SDG File (CSF), as it may be helpful in the assessment.  Verify that the analysis dates on
    the Form 12-IN and the raw data are identical. Review the SDG Narrative and raw data preparation
    logs to determine if samples were properly preserved. If there is an indication of problems with the
    samples, the sample integrity may be compromised. Use professional judgment to evaluate the effect
    of the problem on the sample results.
E.  Action
    NOTE:     Apply the action to each field sample  for which the preservation or holding time criteria
               was not met.
    1.   If the pH of aqueous/water samples is > 2 at the time of sample receipt, determine if the
        laboratory adjusted the pH to < 2 at the time of sample receipt. Also determine if the laboratory
        adjusted the pH to < 2 for the TCLP and SPLP leachates after completion of the leaching
        procedure.  If not, use professional judgment to qualify the samples based on the pH of the
        sample and the chemistry of Mercury (possible Methylation).  Detects should be qualified as
        estimated low (J-) and non-detects as unusable (R).
    2.   If soil/sediment samples are not maintained at < 6°C (but not frozen) from the time of collection
        until receipt at the laboratory, detects should be qualified as estimated low (J-) and non-detects as
        unusable (R).
    3.   If technical holding times are exceeded, use professional judgment to determine the reliability of
        the data based on the  magnitude of the additional time compared to the technical requirement and
        whether the samples were properly preserved. The expected bias would be low. Detects should
        be qualified as estimated low (J-) and non-detects as unusable (R).

August 2014                                                                                  81

-------
Inorganic Data Review
Mercury
    4.  Due to limited information concerning holding times for soil/sediment samples, use professional
       judgment when deciding whether to apply the aqueous/water holding time criteria to
       soil/sediment samples.  If they are applied, document this action in the Data Review Narrative.
    5.  If samples are received with shipping container temperatures > 10°C, use professional judgment
       to determine the reliability of the data, or qualify detects as estimated (J) and non-detects as
       estimated (UJ).
    6.  When shipping or storage temperatures grossly exceed the requirements, the loss of volatile
       mercury compounds or metallic mercury is possible. The expected bias would be low.  Use
       professional judgment to qualify the samples and note it for Regional Laboratory Contracting
       Officer Representative (COR) action.
    7.  When the holding times are exceeded, annotate any possible consequences for the analytical
       results in the Data Review Narrative, and note it for Regional Laboratory COR action.

             Table 22. Preservation and Holding Time Actions for Mercury Analysis
Criteria
Aqueous/water samples received with pH > 2 and pH
not adjusted
TCLP/SPLP leachate samples with pH > 2 and pH not
adjusted
Soil/sediment samples not maintained at < 6°C (but
not frozen) from time of collection until receipt at the
laboratory
Technical Holding Time:
Aqueous/water and TCLP/SPLP leachate samples >
28 days
Technical Holding Time:
Soil/sediment samples > 28 days
Samples received > 10°C*
Action
Detect
Use professional
judgment
J-
Use professional
judgment
J-
J-
J-
J-
Use professional
judgment
J
Non-detect
Use professional
judgment
R
Use professional
judgment
R
R
R
R
Use professional
judgment
UJ
       For samples received with shipping container temperatures > 10°C, Regional policy or project
       Data Quality Objectives (DQO) may allow the use of higher temperature criteria before assessing
       any actions for the affected samples.
August 2014
      82

-------
Inorganic Data Review                                                                     Mercury

                                       II.     Calibration
A.  Review Items
    Form 2-IN, Form 12-IN, Form 15-IN, Form 16-IN, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on initial calibration and
    calibration verification.

C.  Criteria
    1.  Initial Calibration
       The instruments shall be successfully calibrated daily (or once every 24 hours), and each time the
       instrument is set up. The calibration date and time shall be included in the raw data. The
       calibration curve shall be prepared by the same method used to prepare the samples for analysis.
       The curve shall be prepared with the samples that will be analyzed using this calibration curve.
       a.  A blank and at least five calibration standards shall be used to establish the calibration curve.
           At least one of the  calibration standards shall be at or below the Contract Required
           Quantitation Limit (CRQL) but  above the Method Detection Limit (MDL). The calibration
           curve shall be fitted using linear regression or weighted linear regression. The curve may be
           forced through zero.  The calibration curve must have a correlation coefficient > 0.995. The
           calculated percent differences (%Ds) for all of the non-zero standards must fall within ±30%
           of the true value of the standard. The y-intercept of the curve must be less than the CRQL.
    2.  Initial and Continuing Calibration Verification
       The acceptance criteria for the Initial Calibration Verification (ICV) and Continuing Calibration
       Verification (CCV) standards are presented in Table 23. These standards shall be prepared by the
       same method used to prepare the samples for analysis.

         Table 23.  Acceptance Criteria for ICV and CCV Standards for Mercury Analysis
Analytical Method
Cold Vapor AA
Inorganic Analyte
Mercury
ICV/CCV Low Limit
(% of True Value)
85
ICV/CCV High Limit
(% of True Value)
115
       a.  Initial Calibration Verification

           1)  Immediately after the system has been calibrated, the accuracy of the initial calibration
               must be verified and documented by the analysis of an ICV solution(s).  If the ICV
               Percent Recovery (%R) falls outside of the control limits, the analysis should be
               terminated, the problem corrected, the instrument recalibrated, and all affected samples
               reanalyzed.
           2)  Only if the ICV is not available from the United States Environmental Protection Agency
               (EPA), analyses shall be conducted using a certified solution of the analyte from an
               independent commercial standard source, at a concentration level other than that used for
               instrument calibration, but within the calibrated range.
       b.  Continuing Calibration Verification

           1)  To ensure accuracy during the course  of each analytical sequence, the CCV shall be
               analyzed and reported.
           2)  The CCV standard shall be analyzed at a frequency of every hour during an analytical
               sequence. The CCV standard shall also be analyzed at the beginning of the analytical
               sequence, and again after the last analytical sample.

August 2014                                                                                   83

-------
Inorganic Data Review                                                                    Mercury

           3)  The analyte concentration in the CCV standard shall be different than the concentration
               used for the ICV, and a concentration equivalent to the mid level of the calibration curve.
           4)  The same CCV standard solution shall be used throughout the analysis for an SDG.
           5)  The CCV shall be analyzed in the same fashion as an actual sample.  If the %R of the
               CCV was outside of the control limits, the analysis should be terminated, the problem
               corrected, the instrument recalibrated, and all analytical samples analyzed since the last
               compliant CCV reanalyzed.
D.  Evaluation
    1.  Verify that the instrument was calibrated daily (once every 24 hours) and each time the
       instrument was set up, utilizing a blank and at least five calibration standards. Confirm that at
       least one of the calibration standards was analyzed at or below the CRQL, but above the MDL.
       Confirm that calibration standards and samples were prepared at the same time.
    2.  Verify that the ICV and CCV standards were analyzed at the specified frequency and at the
       appropriate concentration. Verify that acceptable %R results were obtained.
    3.  Recalculate one or more of the ICV or CCV %R using the following equation and verify that the
       recalculated value agrees with the laboratory-reported values on Form 2-IN.

                                         Found (value)
                                   %R=	/  ,  / xlOO
                                          True (value)
       Where,
       „    j /  i  x      Concentration (in ug/L) of mercury measured in the analysis of the ICV or
       Found (value)   =  „„,,  i *•
              v    '      CCV solution
       True (value)    =  Concentration (in (ig/L) of mercury in the ICV or CCV source

E.  Action
    NOTES:   For initial calibrations or ICV standards that do not meet the technical criteria,  apply the
               action to the associated samples reported from the analytical sequence.
               For CCV standards that do not meet the technical criteria, apply the action to all samples
               analyzed between a previous technically acceptable analysis of the Quality Control (QC)
               sample and a subsequent technically acceptable analysis of the QC sample in the
               analytical sequence.
    1.  If the instrument was not calibrated daily and each time the instrument was set up, qualify detects
       and non-detects as unusable (R).  If the instrument was not calibrated with at least the minimum
       number of standards, or if the calibration curve does not include standards at  required
       concentrations (e.g., a blank, and at least one standard at or below the CRQL but above the
       MDL), or if the instrument was not calibrated with standards prepared at the  same time as the
       samples, use professional judgment to qualify detects as estimated (J) or unusable (R), and non-
       detects as estimated (UJ) or unusable (R).
    2.  If the correlation coefficient is < 0.995, the %D is outside the ±30% limit, or  the y-intercept is >
       CRQL, qualify detects as estimated (J) and non-detects as estimated (UJ).
    3.  If the ICV or CCV %R falls outside the acceptance windows, use  professional judgment to
       qualify all associated data. If possible, indicate the bias in the review. The following guidelines
       are recommended:
       a.  If the ICV or CCV %R is < 70%, use professional judgment to qualify detects as estimated
           low (J-) or unusable (R), and non-detects as unusable (R).
       b.  If the ICV or CCV %R falls within the range of 70-84%,  qualify detects as estimated low (J-)
           and non-detects as estimated (UJ).

August 2014                                                                                   84

-------
Inorganic Data Review
Mercury
       c.  If the ICV or CCV %R falls within the range of 85-115%, detects and non-detects should not
           be qualified.
       d.  If the ICV or CCV %R falls within the range of 116-130%, qualify detects as estimated high
           (J+). Non-detects should not be qualified.
       e.  If the ICV or CCV %R is > 130%, use professional judgment to qualify detects as estimated
           high (J+) or unusable (R).  Non-detects should not be qualified.
       f  If the ICV or CCV %R is > 165%, qualify detects as unusable (R). Non-detects should not be
           qualified.
    4.  If the laboratory failed to provide adequate calibration information, notify the Regional
       Laboratory COR. The Regional Laboratory COR may contact the laboratory and request the
       necessary information. If the information is unavailable, use professional judgment to assess the
       data.
    5.  Annotate the potential effects on the reported data due to exceeding the calibration criteria in the
       Data Review Narrative.
    6.  If calibration criteria are grossly exceeded, note this for Regional Laboratory COR action.
    NOTE:     For truly critical samples, a further in-depth evaluation of the calibration curve may be
               warranted to determine if additional qualification is necessary.

                       Table 24. Calibration Actions for Mercury Analysis
Criteria
Calibration not performed
Calibration incomplete
Correlation coefficient < 0.995; %D outside
±30%; y-intercept > CRQL
ICV/CCV %R < 70%
ICV/CCV %R 70-84%
ICV/CCV %R 85-1 15%
ICV/CCV %R 116-130%
ICV/CCV %R> 130%
ICV/CCV %R> 165%
Action
Detect
R
Use professional
judgment
JorR
J
Use professional
judgment
J-orR
J-
No qualification
J+
Use professional
judgment
J+orR
R
Non- detect
R
Use professional
judgment
UJorR
UJ
R
UJ
No qualification
No qualification
No qualification
No qualification
August 2014
      85

-------
Inorganic Data Review                                                                    Mercury

                                         III.    Blanks
A.  Review Items
    Form 1-IN, Form 3-IN, Form 12-IN, preparation logs, calibration standard logs, instrument logs, and
    raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the blank responses by
    determining the existence and magnitude of contamination resulting from laboratory (or field)
    activities or baseline drift during analysis.

C.  Criteria
    1.  No contaminants should be found in the blank(s).
    2.  The Initial  Calibration Blank (ICB) shall be analyzed at each mass used for analysis after the
       analytical standards, but not before analysis of the ICV during the initial calibration of the
       instrument (see Section II.C.l). The ICB shall be prepared by the same method used to prepare
       the samples for analysis.
    3.  A Continuing Calibration Blank (CCB) shall be analyzed immediately after every CCV.  The
       CCB shall be prepared by the same method used to prepare the samples for analysis. The CCB
       shall be analyzed at a frequency of every hour during the analytical sequence. The  CCB shall be
       analyzed at the beginning of the analytical sequence, and again after the last CCV that was
       analyzed after the last analytical sample of the analytical sequence. The CCB result (absolute
       value) shall not exceed the CRQL.
    4.  At least one Preparation Blank shall be prepared and analyzed for each matrix, with every SDG,
       or with each batch of samples digested, whichever is more frequent.  The Preparation Blank
       consists of reagent water processed through the  appropriate sample preparation and analysis
       procedure.
    5.  If the analyte concentration in the Preparation Blank is > CRQL, the lowest concentration of the
       analyte in the associated samples must be > lOx the  Preparation Blank concentration.  Otherwise,
       all associated samples with the analyte's concentration < lOx the Preparation Blank
       concentration, and >  CRQL,  should be redigested and reanalyzed.  The laboratory is not to
       correct the  sample concentration for the blank value.
    6.  If the analyte concentration in the Preparation Blank is < (-CRQL), all associated samples with
       the analyte's concentration < lOxthe CRQL, should be redigested and reanalyzed.
    7.  At least one Leachate Extraction Blank (LEB) shall  be prepared and analyzed for each batch of
       samples extracted by TCLP or SPLP. The LEB consists of reagent water processed through the
       extraction procedure. Post-extraction, the LEB  shall be processed through the appropriate sample
       preparation and analysis procedure.
D.  Evaluation
    1.  Verify that an ICB was analyzed after the calibration, the CCB was analyzed at the  specified
       frequency and sequence during the analytical sequence, and Preparation Blanks are prepared and
       analyzed as appropriate for the SDG (e.g., total  number of samples, various types of matrices
       present, number of digestion batches, etc.).
    2.  Review the results reported on Form 3-IN, as well as the raw data for all blanks, and verify that
       the results are accurately reported.
    3.  Evaluate all of the associated blanks for the presence of the target analyte. Verify that  if the
       concentration of the target analyte was > CRQL in a Preparation Blank, all associated samples
       with analyte's concentration > CRQL but < lOx the  Preparation Blank concentration were
       redigested and reanalyzed for that analyte. Verify that if the concentration was < (-CRQL) in a

August 2014                                                                                  86

-------
Inorganic Data Review                                                                   Mercury

       Preparation Blank, all associated samples with the analyte's concentration < lOx CRQL were
       redigested and reanalyzed. Verify that if the absolute value of the target analyte was > CRQL in
       an ICB or a CCB, the analysis was terminated, the problem corrected, the instrument recalibrated,
       and the preceding 10 analytical samples or all analytical samples analyzed since the last
       compliant calibration blank reanalyzed.
E.  Action
    NOTES:  For ICBs that do not meet the technical criteria, apply the action to all associated samples
              reported from the analytical sequence.
              For CCBs that do not meet the technical criteria, apply the action to all associated
              samples analyzed between a previous technically acceptable analysis of the CCB and a
              subsequent technically acceptable analysis of the CCB in the analytical sequence.
              For Preparation Blanks that do not meet the technical criteria, apply the action to all
              associated samples prepared in the same preparation batch. For LEBs that do not meet
              the technical criteria, apply the action to all associated samples  extracted in the same
              extraction batch.
    1.  If the appropriate blanks were not analyzed with the correct frequency, use professional judgment
       to determine if the  associated sample data should be qualified; obtain additional information from
       the laboratory, if necessary. Record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action.
    2.  Action regarding unsuitable blank results depends on the circumstances and origin of the blank.
       In instances where more than one blank is associated with a given sample, qualification should be
       based upon a comparison with the associated blank having the highest concentration of
       contaminant.
    3.  Some general "technical" review actions include:
       a.   For any blank (including Preparation Blanks and LEBs) reported with detects < CRQL, report
           detects < CRQL at the CRQL and qualify as non-detect (U).  For any blank (including
           Preparation Blanks and LEBs) reported with a detect < CRQL, use professional judgment to
           qualify the sample results > CRQL.  Non-detects should not be qualified.
       b.   For any blank (including Preparation Blanks and LEBs) reported with a negative result
           < (-MDL) but > (- CRQL), carefully evaluate it to determine its effect on the sample data.
           Use professional judgment to assess the data.
       c.   The blank analyses may not involve the same weights, volumes, or dilution factors as the
           associated samples. In particular, soil/sediment sample results reported on Form 1-IN will not
           be on the same basis (units, dilution) as the calibration blank data reported on Form 3-IN.  It
           may be easier to work with the raw data  and/or convert the ICB or CCB results to the same
           units as the soil/sediment samples for comparison purposes.
    4.  Specific "method" actions include:
       a.   If an ICB or a CCB result is > CRQL, the analysis should be terminated.  If the analysis was
           not terminated and the associated samples were not reanalyzed, non-detects should not be
           qualified.  Report detects < CRQL at CRQL and qualify as non-detect (U).  Report sample
           results that are > CRQL but < ICB/CCB  Results at ICB/CCB Results and use professional
          judgment to qualify as non-detect (U) or unusable (R). Use professional judgment to qualify
           sample results > ICB/CCB Results. Record the situation in the Data Review Narrative, and
           note it for Regional Laboratory COR action.
August 2014                                                                                  87

-------
Inorganic Data Review                                                                   Mercury

       b.  If an ICB or a CCB result is < (-CRQL), the analysis should be terminated.  If the analysis
           was not terminated and the associated samples were not reanalyzed, use professional
           judgment to qualify non-detects as estimated (UJ) or unusable (R). Use professional
           judgment to qualify detects < CRQL or qualify as estimated low (J-). Use professional
           judgment to qualify sample results that are > CRQLs as estimated low (J-).
       c.  If the concentration of the analyte in the Preparation Blank/LEB is > CRQL, the lowest
           concentration of that analyte in the associated samples must be > lOx the Preparation
           Blank/LEB concentration.  All samples associated with the Preparation Blank with
           concentrations < 1 Ox the Preparation Blank concentration and >  CRQL should have been
           redigested and reanalyzed.  If the associated samples were not redigested and reanalyzed,
           report the sample results at Preparation Blank Results; use professional judgment to qualify
           as estimated high (J+) or unusable (R).  Report results <10x the LEB concentration and >
           CRQL in the samples associated with the LEB at LEB Results; use professional judgment to
           qualify the results as estimated high (J+) or unusable (R). Report detects < CRQLs in the
           samples associated with the Preparation Blank/LEB at CRQLs and qualify as non-detect (U).
           Non-detects and sample results that are > lOx Preparation Blank/LEB Results should not be
           qualified.  If the laboratory failed to redigest and reanalyze the samples associated with the
           Preparation Blank, record it in the Data Review Narrative, and note it for Regional
           Laboratory COR action.
       d.  For any Preparation Blank or LEB reported with a negative result, < (-CRQL), use
           professional judgment to qualify detects < CRQL or qualify as estimated low (J-). Qualify
           sample results that are > CRQLs as estimated low (J-), and qualify non-detects as estimated
           (UJ).  Sample results that are > lOx CRQLs  should not be qualified.
August 2014                                                                                 88

-------
Inorganic Data Review
Mercury
                        Table 25. Blank Actions for Mercury Analysis
Blank
Type
ICB/CCB
ICB/CCB
ICB/CCB
ICB/CCB
Preparation
Blank/LEB
Preparation
Blank/LEB
Preparation
Blank/LEB
Preparation
Blank/LEB
Blank Result
Detect < CRQL
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Detect < CRQL
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Sample Result
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but
< ICB/CCB Result
> ICB/CCB Result
Non-detect
Detect < CRQL
>CRQL
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but
< 1 Ox the Preparation
Blank/LEB Result
> lOx the Preparation
Blank/LEB Result
Non-detect
Detect < CRQL
< lOx CRQL
> lOx CRQL
Action
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at ICB/CCB Result as non-
detect (U) or unusable (R)
Use professional judgment
Use professional judgment to qualify
as estimated (UJ) or unusable (R)
Use professional judgment or (J-)
Use professional judgment to qualify
as estimated low (J-)
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at Preparation Blank/LEB
Result and use professional judgment
to qualify results as estimated high
(J+) or unusable (R)
No qualification
Qualify as estimated (UJ)
Use professional judgment or (J-)
Report results > CRQL as estimated
low (J-)
No qualification
August 2014
     89

-------
Inorganic Data Review                                                                  Mercury

                               IV.    Duplicate Sample Analysis
A.  Review Items
    Cover Page, Form 6-IN, instrument printouts, and raw data.

B.  Objective
    The objective of duplicate sample analysis is to demonstrate acceptable method precision by the
    laboratory at the time of analysis.

C.  Criteria
    1.   Samples identified as field blanks or Performance Evaluation (PE) samples cannot be used for
        duplicate sample analysis.
    2.   At least one duplicate sample shall be prepared and analyzed from each group of samples of a
        similar matrix type (e.g., water or soil) or for each SDG. Duplicates cannot be averaged for
        reporting on Form 1-IN.  Additional duplicate sample analyses may be required by EPA Regional
        request.  Alternately, the Region may require that a specific sample be used for the duplicate
        sample analysis.
    3.   A control limit of 20% for the Relative Percent Difference (RPD) shall be used for original  and
        duplicate sample values > 5x the CRQL.
    4.   A control limit of the CRQL shall be used if either the sample or duplicate value is < 5x the
        CRQL.  The absolute value of the control limit  (CRQL)  shall be entered in the "Control Limit"
        column on Form 6-IN. If both samples are non-detects, the RPD is not calculated for Form 6-IN.
    NOTE:     The above control limits are method requirements for duplicate samples, regardless of
               the sample matrix type.  However, it should be noted that laboratory variability arising
               from the sub-sampling of non-homogenous soil  samples is a common occurrence.
               Therefore, for technical review purposes only, Regional policy or project DQOs may
               allow the use of less restrictive criteria  (e.g., 35% RPD, 2x the CRQL) to be assessed
               against duplicate soil samples.
D.  Evaluation
    1.   Verify, from the Cover Page and the raw data, that the appropriate number of required duplicate
        samples were prepared and analyzed for the SDG.
    2.   Verify, using Form 6-IN and the raw data, that the duplicate results fall within the established
        control limits.
    3.   Verify that a field blank or PE sample was not used for duplicate analysis.
    4.   Check the raw data and recalculate one or more of the RPD values using the following equation
        to verify that the results were correctly reported on Form 6-IN:
                                             IS-DI
                                    RPD= —•	— x 100
                                           (S + D)/2

        Where,
        RPD   = Relative Percent Difference
        S      =  Sample Result (original)
        D      = Duplicate Result
August 2014                                                                                 90

-------
Inorganic Data Review                                                                   Mercury

E.  Action
    NOTE:     For a duplicate sample analysis that does not meet the technical criteria, apply the action
               to all samples of the same matrix if the samples are considered sufficiently similar.
               Exercise professional judgment in determining sample similarity when making use of all
               available data, including: site and sampling documentation (e.g., location and type of
               sample, descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity,
               chlorine); and laboratory data for other parameters [e.g., Total Suspended Solids (TSS),
               Total Dissolved Solids (TDS), Total Organic Carbon (TOC), alkalinity or buffering
               capacity, reactive sulfide, anions].  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG. Two
               determinations are: 1) only some samples in the SDG are similar to the duplicate sample,
               and that only these samples should be qualified; or 2) no samples are sufficiently similar
               to the sample used for the duplicate, and thus only the field sample used to prepare the
               duplicate sample should be qualified.
    1.  If the appropriate number of duplicate samples was not analyzed for each matrix using the correct
       frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary.  Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Associated samples
       that are detects should be qualified as estimated (J) and non-detects as estimated (UJ) if any of the
       frequency criteria is not met.
    2.  If both original sample and duplicate sample results are > 5x the CRQL and the RPD is > 20%,
       qualify detects as estimated (J), and non-detects as estimated (UJ).

    3.  If both original sample and duplicate sample results are > 5x the CRQL and the RPD is < 20%,
       detects and non-detects should not be qualified.
    4.  If RPD > 100%, use professional judgment to determine if the associated  sample data should be
       qualified.
    5.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate > CRQL, qualify detects as estimated (J), and
       non-detects as estimated (UJ).
    6.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate < CRQL, detects and non-detects should not be
       qualified.
    7.  If a field blank or PE sample was used for the duplicate sample  analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be  carefully checked.  Exercise
       professional judgment when evaluating the data.
    8.  Annotate the potential effects on the data due to out-of-control duplicate sample results in the
       Data Review Narrative.
August 2014                                                                                  91

-------
Inorganic Data Review
Mercury
                   Table 26. Duplicate Sample Actions for Mercury Analysis
Criteria
Both original sample and duplicate sample results are > 5x
the CRQL and RPD > 20%*
Both original sample and duplicate sample results are > 5x
the CRQL and RPD is < 20%
RPD > 100%
Original sample or duplicate sample results < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate > CRQL*
Original sample or duplicate sample result < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate < CRQL
Action
Detect
J
No qualification
Use professional
judgment
J
No qualification
Non-detect
UJ
No qualification
Use professional
judgment
UJ
No qualification
    *   The above control limits are method requirements for duplicate samples, regardless of the
       sample matrix type.  However, it should be noted that laboratory variability arising from the sub-
       sampling of non-homogenous soil samples is a common occurrence. Therefore, for technical
       review purposes only, Regional policy or project DQOs may allow the use of less restrictive
       criteria (e.g., 35% RPD, 2xthe CRQL) to be assessed against duplicate soil samples.
August 2014
      92

-------
Inorganic Data Review                                                                   Mercury

                                 V.     Spike Sample Analysis
A.  Review Items
    Cover Page, Form 5A-IN, instrument printouts, and raw data.

B.  Objective
    The objective of the spiked sample analysis is to  evaluate the effect of each sample matrix on the
    sample preparation procedures and the measurement methodology.

C.  Criteria
    1.   Samples identified as field blanks or PE samples cannot be used for spiked sample analysis.
    2.   At least one spiked sample shall be prepared and analyzed from each group of samples with a
        similar matrix type (e.g., water or soil), or for each SDG.

    3.   The spike %R shall be within the established acceptance limits. However, spike recovery limits
        do not apply when the sample concentration  is > 4x the spike added. In such an event, the data
        shall be reported unflagged, even if the %R does not meet the acceptance criteria.
    4.   If the spiked sample analysis was performed  on the same sample that was chosen for the duplicate
        sample analysis, spike calculations shall be performed using the results of the sample designated
        as the "original sample."  The average of the duplicate results cannot be used for the purpose of
        determining %R.

    NOTE:     The final spike concentration required is presented in the method described in the SOW.
D.  Evaluation
    1.   Verify, using the Cover Page, Form 5 A-IN and raw data, that the appropriate number of required
        spiked samples was prepared and analyzed for the SDG.
    2.   Verify that a field blank or PE sample was not used for the spiked sample analysis.
    3.   Verify, using Form 5 A-IN and the raw data, that all Matrix Spike sample results fall within the
        established control limits.
    4.   Recalculate, using the raw data, one or more  of the %Rs using the following equation, and verify
        that the recalculated value agrees with the laboratory-reported values on Form 5A-IN:


                                                   SSR-SR
                                      %Recovery = ———  x  100
                                                      oA.
        Where,
        SSR    =  Spiked Sample Result
        SR     =  Sample Result
        SA     =  Spike Added

    NOTE:     When the sample result is < MDL or reported as a non-detect,  use SR = 0 only for the
               purpose of calculating the %R. The  actual spiked sample result, sample result, and %R
               (positive or negative) shall still be reported on Forms 5A-IN.
E.  Action
    NOTE:     For a Matrix Spike that does not meet the technical criteria, apply the action to all
               samples of the same matrix if the samples are considered sufficiently similar.  Exercise
               professional judgment in determining sample similarity when making use of all available
               data, including: site and sampling documentation (e.g., location and type of sample,
               descriptive data, soil classification);  field test data (e.g., pH, Eh, conductivity, chlorine);
               and laboratory data for other parameters (e.g., TSS, TDS, TOC, alkalinity or buffering
               capacity, reactive sulfide, anions). Additionally, use the sample data (e.g., similar

August 2014                                                                                   93

-------
Inorganic Data Review
Mercury
               concentrations of analytes) in determining similarity between samples in the SDG.
               Possible determinations are:  1) only some of the samples in the SDG are similar to the
               Matrix Spike sample, and that only these samples should be qualified; or, 2) no samples
               are sufficiently similar to the sample used for the Matrix Spike, and thus only the field
               sample used to prepare the Matrix Spike sample should be qualified.
    1.  If the appropriate number of Matrix Spike samples was not analyzed for each matrix using the
       correct frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary.  Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If a field blank or PE sample was used for the spiked sample analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be  carefully checked.  Use
       professional judgment when evaluating the data. Detects should be qualified as estimated (J) and
       non-detects as estimated (UJ).
    3.  If the Matrix Spike %R is < 30%, qualify detects as estimated low (J-) and non-detects as
       unusable (R).
    4.  If the Matrix Spike %R falls within the range of 30-74%, qualify detects as estimated low (J-) and
       non-detects as estimated (UJ).
    5.  If the Matrix Spike %R falls with the range of 75-125%, detects and non-detects should not be
       qualified.
    6.  If the Matrix Spike %R is > 125%, qualify detects as estimated  high (J+).  Non-detects should not
       be qualified.
    7.  Annotate the potential effects on the data due to out-of-control spiked sample results in the Data
       Review Narrative.

                     Table 27. Spike Sample Actions for Mercury Analysis
Criteria
Matrix Spike %R < 30%
Matrix Spike %R 30-74%
Matrix Spike %R75-125%
Matrix Spike %R > 125%
Action
Detect
J-
J-
No qualification
J+
Non- detect
R
UJ
No qualification
No qualification
    NOTE:     The above control limits are method requirements for spike samples, regardless of the
               sample matrix type. However, it should be noted that laboratory variability arising from
               the sub-sampling of non-homogenous soil samples is a common occurrence. Therefore,
               for technical review purposes only, Regional policy or project DQOs may allow the use
               of less restrictive criteria (e.g., 10 %R and 150 %R for the lower and upper limits) to be
               assessed against spike soil samples.
August 2014
      94

-------
Inorganic Data Review                                                                   Mercury

                     VI.     Regional Quality Assurance and Quality Control
A.  Review Items
    Form 1 -IN, instrument printouts, and raw data.

B.  Objective
    The objective is to use results from the analysis of Regional Quality Assurance/Quality Control
    (QA/QC) samples such as field blanks, PE samples, blind spikes, and blind blanks to determine the
    validity of the analytical results.
C.  Criteria
    Criteria are determined by the Region.
D.  Evaluation
    Evaluation procedures must follow the Region's Standard Operating Procedure (SOP) for data
    review. Each Region will handle the evaluation of PE samples on an individual basis. Compare
    results for PE samples with the acceptance criteria for the specific PE samples if possible.
    Calculate the RPD between field duplicates and provide his information in the Data Review
    Narrative.
E.  Action
    Any action must be in accordance with Regional specifications and criteria for acceptable PE sample
    results. Note any unacceptable PE sample results for Regional Laboratory COR action.
August 2014                                                                                  95

-------
Inorganic Data Review                                                                     Mercury

                               VII.    Overall Assessment of Data
A.  Review Items
    Entire sample data package, data review results, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data (including any confirmation data).

B.  Objective
    The objective is to provide the overall assessment on data quality and usability.

C.  Criteria
    1.  Review all available materials to assess the overall quality of the data, keeping in mind the
       additive nature of analytical problems.
    2.  Reported analyte concentrations must be quantitated according to the appropriate analytical
       method, as listed in the method. All sample results must be within the linear calibration ranges
       per methods. Percent Solids (%Solids) must be properly used for all applicable matrix result
       calculations.
D.  Evaluation
    Examine the raw data to verify that the correct calculation of the sample results was reported by the
    laboratory. Digestion logs, instrument printouts, etc., should be compared to the reported sample
    results recorded on the appropriate Inorganic Summary Forms (Form 1-IN through Form 16-IN).
    1.  Evaluate any technical problems not previously addressed.
    2.  Examine the raw data for any anomalies (e.g., baseline shifts, negative absorbance, omissions,
       illegibility, etc.).
    3.  Verify that the appropriate methods and amounts were used to prepare samples and standards for
       analysis. If reduced volumes are used, verify that the laboratory received Regional Laboratory
       COR approval for the use of the reduced volume.
    4.  Verify that there are no transcription or reduction errors (e.g., dilutions, %Solids, sample weights,
       etc.) on one or more samples. Recalculate %Solids for at least 10% of the samples and verify that
       the calculated %Solids agree with that reported by the laboratory.
    5.  Verify that the MDL is properly reported and that it is not greater than the CRQL.

    6.  Verify that results fall within the calibrated range (Form 15-IN).
    7.  If appropriate information is available, assess the usability of the data to assist the data user in
       avoiding inappropriate use of the data.  Review all available information,  including the Quality
       Assurance Project Plan (QAPP), focusing specifically on the acceptance or performance criteria,
       the SOPs, and communication with the user concerning the intended use and desired quality of
       these data.
E.  Action
    1.  Use professional judgment to determine if there is any need to qualify data which are not
       qualified based on the QC criteria previously discussed.
    2.  Use professional judgment to qualify detects and non-detects  if the MDL  exceeds CRQL.
    3.  If a sample is not diluted properly when sample results exceed the upper limit of the calibration
       range, qualify detects as estimated (J).
    4.  Write a brief Data Review Narrative to give the user an indication of the analytical limitations of
       the data. Annotate any discrepancies between the data and the SDG Narrative for Regional
       Laboratory COR action. If sufficient information on the intended use and required quality of the
       data is available, include an assessment of the data usability within the given context.
August 2014                                                                                    96

-------
Inorganic Data Review                                                                     Mercury

    5.  If any discrepancies are found, notify the Regional Laboratory COR. The Regional Laboratory
       COR may contact the laboratory to obtain additional information for resolution. If a discrepancy
       remains unresolved, use professional judgment to determine if qualification of the data is
       warranted.
August 2014                                                                                   97

-------
Inorganic Data Review                                                                  Mercury

                                     VIII.   Calculations
Aqueous/Water Samples:
                              Hg Concentration (ug/L) = C x DF
    Where,
    C      =  Instrument value in |o,g/L from the calibration curve
    DF     =  Dilution Factor of the original sample
Soil/Sediment Samples:
                  Hg Concentration (mg/kg dry weight) = C x ——- x DF x 0.1
                                                          W x S
    Where,
    C      =  Instrument value in |o,g/L from the calibration curve
    W     =  Initial aliquot amount (g)
    S      =  %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    =  Dilution Factor
Adjusted MDL/Adjusted CRQL Calculation:
    To calculate the adjusted MDL or adjusted CRQL for aqueous/water samples, substitute the value of
    the MDL ((ig/L) or CRQL ((ig/L) into the "C" term in the equation above.
    Calculate the adjusted MDL or adjusted CRQL for soil/sediment samples as follows:

                                                            Wm
                      Adjusted MDL or CRQL (mg/kg) = C x  ——- x DF
                                                           W x S

    Where,
    C      =  MDL or CRQL (mg/kg)
    Wm    =  Method required minimum sample weight (g) (0.50 g)
    W     =  Initial aliquot amount (g)
    S      =  %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    =  Dilution Factor
August 2014                                                                                98

-------
Inorganic Data Review                                                                  Cyanide

                                 CYANIDE DATA REVIEW
The inorganic data requirements for cyanide to be reviewed during validation are listed below:
Example Analytical Sequence	101
I.    Preservation and Holding Times	103
II.   Calibration	105
III.  Blanks	109
IV.  Duplicate Sample Analysis	113
V.   Spike Sample Analysis	116
VI.  Regional Quality Assurance and Quality Control	119
VII.  Overall Assessment of Data	120
VIII. Calculations	122
August 2014                                                                                 99

-------
Inorganic Data Review                                                                     Cyanide
                                This page is intentionally left blank.
August 2014                                                                                   100

-------
Inorganic Data Review                                                             Cyanide

                              Example Analytical Sequence
The following is an example of an analytical sequence:

S##
S##
s##
s##
s##
s##
ICV
ICB
ccv###
CCB###
samples
ccv###
CCB###
samples
ccv###
CCB###, etc.
* Suffix ## and ### are as specified in Exhibit B of the Statement of Work (SOW).
August 2014                                                                         101

-------
Inorganic Data Review                                                                     Cyanide
                                This page is intentionally left blank.
August 2014                                                                                   102

-------
Inorganic Data Review                                                                    Cyanide

                              I.  Preservation and Holding Times
A.  Review Items
    Form 1-IN, Form 12-IN, Traffic Report/Chain of Custody (TR/COC) documentation, Form DC-1,
    raw data, and the Sample Delivery Group (SDG) Narrative checking for: pH; shipping container
    temperature; holding time; and other sample conditions.

B.  Objective
    The objective is to determine the validity of the analytical results based on sample conditions and the
    technical holding time of the sample.

C.  Criteria
    1.   The technical holding time is determined from the date of collection, or the date Synthetic
        Precipitation Leaching Procedure (SPLP) extraction is complete, to the date of analysis.
    2.   The technical holding time criteria for aqueous/water samples and leachate samples from SPLP is
        14 days,  preserved (with sodium hydroxide) to pH > 10.
    3.   The technical holding time criteria for soil/sediment samples is  14 days, based on the technical
        holding time criteria for aqueous/water samples.

    4.   Aqueous/water and soil/sediment samples shall be maintained at < 6°C (but not frozen) from the
        time of collection until preparation. The SPLP leachates must be stored at < 6°C (but not frozen)
        from the time of the leaching procedure completion until preparation.

D.  Evaluation
    Establish technical holding times by comparing the sampling date(s) on the TR/COC documentation
    with the dates of analysis on Form 12-IN and the raw data; also consider using information in the
    Complete SDG File (CSF), as it may be helpful in the assessment.  Verify that the analysis dates on
    the Form 12-IN  and the raw data are identical. Review the SDG Narrative and raw data preparation
    logs to determine if samples were properly preserved. If there is an indication of problems with the
    samples, the  sample integrity may be compromised. Use professional judgment to evaluate the effect
    of the problem on the sample results.  For aqueous/water samples, look for evidence that the samples
    were tested for the presence of sulfides, oxidizing agents, or nitrate/nitrite, and whether the
    appropriate preservation steps were taken.
E.  Action
    NOTE:    Apply the action to each field sample for which the preservation or holding time criteria
              were not met.
    1.   If oxidizing  agents were detected in aqueous/water samples at the time of sample preparation,
        qualify detects as estimated low (J-) and non-detects as unusable (R). If sulfides were detected in
        aqueous/water samples at the time of sample preparation, qualify detects as estimated (J) and
        non-detects  as unusable (R).  If there is evidence that samples were not treated with sulfamic acid
        prior to distillation for nitrate/nitrite interferences, qualify detects as estimated (J) and non-detects
        as unusable  (R).  If the pH of aqueous/water samples was <  10 at the time of sample receipt, use
        professional judgment to qualify the samples based on the pH of the sample. Detects should be
        qualified as  estimated low (J-) and non-detects as unusable (R).
    2.   If aqueous/water and soil/sediment samples are not maintained at < 6°C (but not frozen) from the
        time of collection until receipt at the laboratory, detects should be qualified as estimated low (J-)
        and non-detects as unusable (R).
    3.   If technical holding times are exceeded, use professional judgment to determine the reliability of
        the data based on the magnitude of the additional time compared to the technical requirement and
        whether the  samples are properly preserved. The expected bias would be low.  Detects should be
        qualified as  estimated low (J-) and non-detects as unusable (R).

August 2014                                                                                 103

-------
Inorganic Data Review
Cyanide
    4.  Due to limited information concerning holding times for soil/sediment samples, use professional
       judgment in deciding whether to apply the aqueous/water holding time criteria to soil/sediment
       samples. If they are applied, document this in the Data Review Narrative.
    5.  When the holding times are exceeded, annotate any possible consequences for the analytical
       results in the Data Review Narrative, and note it for Regional Laboratory Contracting Officer
       Representative (COR) action.

             Table 28. Preservation and Holding Time Actions for Cyanide Analysis
Criteria
Aqueous/water samples received with oxidizing agents
present
Aqueous/water samples received with sulfides present
Aqueous/water samples received with nitrate present and not
treated with sulfamic acid
Aqueous/water samples received with pH < 10
Aqueous/water and soil/sediment samples not maintained at
< 6°C (but not frozen) from time of collection until receipt at
the laboratory
Technical holding time:
Aqueous/water and SPLP leachate samples > 14 days
Technical holding time:
Soil/sediment samples > 14 days
Action
Detect
J-
J
J
Use professional
judgment
J-
Use professional
judgment
J-
J-
J-
Non- detect
R
R
R
Use professional
judgment
R
Use professional
judgment
R
R
R
August 2014
    104

-------
Inorganic Data Review                                                                     Cyanide

                                       II.     Calibration
A.  Review Items
    Form 2-IN, Form 12-IN, Form 15-IN, Form 16-IN, preparation logs, calibration standard logs,
    instrument logs, instrument printouts, and raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on initial calibration and
    calibration verification.

C.  Criteria
    1.  Initial Calibration
       The instruments shall be successfully calibrated daily (or once every 24 hours), and each time the
       instrument is set up. The calibration date and time shall be included in the raw data. The
       calibration curve standards shall be distilled by the same method used to prepare the samples for
       analysis.
       a.  A blank and at least five calibration standards shall be employed to establish the calibration
           curve. At least one of the calibration standards shall be at or below the Contract Required
           Quantitation Limit (CRQL), but above the Method Detection Limit (MDL).  The calibration
           curve shall be fitted using linear regression or weighted linear regression. The curve may be
           forced through zero.  The calibration curve must have a correlation coefficient > 0.995. The
           calculated percent differences (%Ds) for all of the non-zero standards must be within ±30%
           of the true value of the standard.  The y-intercept of the curve must be less than the CRQL.
    2.  Initial and Continuing Calibration Verification
       The acceptance criteria for the Initial Calibration  Verification (ICV) and Continuing Calibration
       Verification (CCV) standards are presented in Table 29:

         Table 29. Acceptance Criteria for ICV and CCV Standards for Cyanide Analysis
Analytical Method
Colorimetric
Inorganic Analyte
Cyanide
ICV/CCV Low Limit
(% of True Value)
85
ICV/CCV High Limit
(% of True Value)
115
       a.  Initial Calibration Verification

           1)  Immediately after each colorimetric system has been calibrated, the accuracy of the initial
               calibration must be verified and documented by the analysis of an ICV solution(s). If the
               ICV Percent Recovery (%R) falls outside of the control limits, the analysis should be
               terminated, the problem corrected, the instrument recalibrated, and all affected samples
               reanalyzed.
           2)  Only if the ICV is not available from the United States Environmental Protection Agency
               (EPA), analyses shall be conducted using a certified solution of the analyte from an
               independent commercial standard source, at a concentration level other than that used for
               instrument calibration, but within the calibrated range.
           3)  The ICV standard solution shall be distilled by the same method used to prepare the
               samples for analysis.
       b.  Continuing Calibration Verification
           1)  To ensure  accuracy during the course of each analytical sequence, the CCV shall be
               analyzed and reported.
August 2014                                                                                   105

-------
Inorganic Data Review                                                                    Cyanide

           2)  The CCV standard shall be analyzed at a frequency of every hour during an analytical
               sequence.  The CCV standard shall also be analyzed at the beginning of the analytical
               sequence, and again after the last analytical sample.
           3)  The analyte concentration in the CCV standard shall be different from the concentration
               used for the ICV, and at a concentration equivalent to the mid level of the calibration
               curve.
           4)  The same CCV standard solution shall be used throughout the analysis for an SDG.
           5)  The CCV shall be processed and analyzed in the same fashion as an actual sample. If the
               %R of the  CCV was outside of the control limits, the analysis should be terminated, the
               problem corrected, the instrument recalibrated, and all analytical samples analyzed since
               the last compliant CCV reanalyzed.
           6)  The CCV standard solution shall be distilled by the same method used to prepare the
               samples for analysis.
D.  Evaluation
    1.  Verify that the instrument was calibrated daily (once every 24 hours) and each time the
       instrument was set up, utilizing a blank and at least five calibration standards. Confirm that at
       least one of the calibration standards was analyzed at or below the CRQL, but above the MDL.
    2.  Verify, using the distillation log, that the calibration standards, the ICV, and the CCV standards
       were distilled and analyzed.
    3.  Verify that the ICV and CCV  standards were analyzed at the specified frequency and at the
       appropriate concentration. Verify that acceptable %R results were obtained.
    4.  Recalculate one or more of the ICV or CCV %R using the following equation and verify that the
       recalculated value agrees with the  laboratory-reported values on Form 2-IN.

                                         Found (value)
                                    %R =	 x  100
                                          True (value)
       Where,
       „    j /  i  x      Concentration (in ug/L) of cyanide measured in the analysis of the ICV or
       Found (value)   = „„,,  i  *•
              v    '      CCV solution
       True  (value)    = Concentration (in (ig/L) of cyanide in the ICV or CCV source
E.  Action
    NOTES:    For initial calibrations or ICV standards that do not meet the technical criteria, apply the
               action to all associated samples reported from the analytical sequence.
               For CCV standards that do not meet the technical criteria, apply the action to all samples
               analyzed between a previous technically acceptable analysis of the  Quality Control (QC)
               sample and a subsequent technically acceptable analysis of the QC  sample in the
               analytical sequence.
    1.  If the instrument was not calibrated daily and each time the instrument was  set up, qualify detects
       and non-detects as unusable (R). If the instrument was not calibrated with at least the minimum
       number of standards, or if the  calibration curve does not include standards at required
       concentrations (e.g., a blank and a standard at or below the CRQL but above the MDL), use
       professional judgment to qualify detects as estimated (J) or unusable (R), and non-detects as
       estimated (UJ) or unusable (R).
    2.  If the correlation coefficient is < 0.995, the %D is outside the ±30% limit, or y-intercept > CRQL,
       qualify detects as estimated (J) and non-detects as estimated (UJ).
August 2014                                                                                  106

-------
Inorganic Data Review                                                                     Cyanide

    3.  If the ICV or the CCV standards are not distilled, qualify detects as estimated (J) and non-detects
       as estimated low (UJ-).
    4.  If the ICV or CCV %R falls outside the acceptance windows, use professional judgment to
       qualify all associated data.  If possible, indicate the bias in the review.  The following guidelines
       are recommended:
       a.   If the ICV or CCV %R is < 70%, qualify detects as estimated low (J-) or unusable (R) and
           non-detects as unusable (R).
       b.  If the ICV or CCV %R falls within the range of 70-84%, qualify detects as estimated  low (J-)
           and non-detects as estimated (UJ).
       c.   If the ICV or CCV %R falls within the range of 85-115%, detects and non-detects should not
           be qualified.
       d.  If the ICV or CCV %R falls within the range of 116-130%, qualify detects as estimated high
           (J+). Non-detects  should not be qualified.
       e.   If the ICV or CCV %R is > 130%, use professional judgment to qualify detects as estimated
           high (J+) or unusable (R).  Non-detects should not be qualified.
    5.  If the laboratory failed to provide adequate calibration information, notify the Regional
       Laboratory COR. The Regional Laboratory COR may contact the laboratory and request  the
       necessary information. If the information is not available, use professional judgment to assess the
       data.
    6.  Annotate the potential effects on the reported data due to exceeding the calibration criteria in the
       Data Review Narrative.
    7.  If calibration criteria are grossly exceeded, note this for Regional Laboratory COR action.
    NOTE:    For truly critical samples, a further in-depth evaluation of the calibration curve may be
               warranted to determine if additional qualification is necessary.
August 2014                                                                                  107

-------
Inorganic Data Review
Cyanide
                      Table 30. Calibration Actions for Cyanide Analysis
Criteria
Calibration not performed
Calibration incomplete
Correlation coefficient < 0.995; %D outside ±30%;
y-intercept > CRQL
Standards and QC not distilled
ICV/CCV %R < 70%
ICV/CCV %R 70-84%
ICV/CCV %R 85-1 15%
ICV/CCV %R 116-130%
ICV/CCV %R> 130%
Action
Detect
R
Use professional
judgment
JorR
J
J
Use professional
judgment
J-orR
J-
No qualification
J+
Use professional
judgment
J+orR
Non-detect
R
Use professional
judgment
UJorR
UJ
UJ-
R
UJ
No qualification
No qualification
No qualification
August 2014
    108

-------
Inorganic Data Review                                                                     Cyanide

                                         III.    Blanks
A.  Review Items
    Form 1-IN, Form 3-IN, Form 12-IN, preparation logs, calibration standard logs, instrument logs, and
    raw data.

B.  Objective
    The objective is to determine the validity of the analytical results based on the blank responses by
    determining the existence and magnitude of contamination resulting from laboratory (or field)
    activities or baseline drift during analysis.

C.  Criteria
    1.  No contaminants should be found in the blank(s).
    2.  The Initial Calibration Blank (ICB) shall be analyzed at each mass used for analysis after the
       analytical standards, but not before analysis of the ICV during the initial calibration of the
       instrument (see Section II.C.l).
    3.  A Continuing Calibration Blank (CCB) shall be analyzed immediately after every CCV.  The
       CCB shall be analyzed at a frequency of every hour during the analytical sequence.  The CCB
       shall be analyzed at the beginning of the analytical sequence, and again  after the last CCV that
       was analyzed after the last analytical sample of the analytical sequence.  The CCB result
       (absolute value) shall not exceed the CRQL.
    4.  At least one Preparation Blank shall be prepared and analyzed for each matrix, with every SDG,
       or with each batch of samples distilled, whichever is more frequent.  The Preparation Blank
       consists of reagent water processed through the  appropriate sample preparation and analysis
       procedure.
    5.  If the analyte concentration in the Preparation Blank is > CRQL, the lowest concentration of the
       analyte in the associated samples must be > lOx the Preparation Blank concentration. Otherwise,
       all associated samples with the analyte's concentration < lOx the Preparation Blank
       concentration, and >  CRQL, should be redistilled and reanalyzed.  The laboratory is not to correct
       the sample concentration for the blank value.
    6.  If the analyte concentration in the Preparation Blank is < (-CRQL), all associated samples with
       the analyte's concentration < lOxthe CRQL, should be redistilled and reanalyzed.
    7.  At least one Leachate Extraction Blank (LEB) shall be prepared and analyzed for each batch of
       samples extracted by SPLP.  The LEB consists of reagent water processed through the extraction
       procedure. Post-extraction, the  LEB shall be processed through the appropriate sample
       preparation and analysis procedure.
D.  Evaluation
    1.  Verify that an ICB was analyzed after the calibration, the CCB was analyzed at the specified
       frequency and sequence during the analytical sequence, and Preparation Blanks and LEBs are
       prepared and analyzed as appropriate for the SDG (e.g., total number of samples, various types of
       matrices present, number of distillation batches, etc.).
    2.  Review the results reported Form 3-IN, as well as the raw data for all blanks, and verify that the
       results were accurately reported.
    3.  Evaluate all of the  associated blanks for the presence of the target analyte. Verify that if the
       concentration of the target analyte was > CRQL in a Preparation Blank,  all associated samples
       with the analyte's concentration > CRQL but <  lOx the Preparation Blank concentration  were
       redistilled and reanalyzed for the analyte.  Verify that if a concentration  was < (-CRQL) in a
       Preparation Blank, all associated samples with the analyte's concentration < lOx CRQL were
       redistilled and reanalyzed. Verify that if the absolute value of the target analyte was > CRQL in

August 2014                                                                                  109

-------
Inorganic Data Review                                                                   Cyanide

       an ICB or a CCB, the analysis was terminated, the problem corrected, the instrument recalibrated,
       and the preceding 10 analytical samples or all analytical samples analyzed since the last
       compliant calibration blank reanalyzed.
E.  Action
    NOTES:  For ICBs that do not meet the technical criteria, apply the action to all associated samples
              reported from the analytical sequence.
              For CCBs that do not meet the technical criteria, apply the action to all associated
              samples analyzed between a previous technically acceptable analysis of the CCB and a
              subsequent technically acceptable analysis of the CCB in the analytical sequence.
              For Preparation Blanks that do not meet the technical criteria, apply the action to all
              associated samples prepared in the same preparation batch. For LEBs that do not meet
              the technical criteria, apply the action to all associated samples extracted in the same
              extraction batch.
    1.  If the appropriate blanks were not analyzed with the correct frequency, use professional judgment
       to determine if the associated sample data should be qualified; obtain additional information from
       the laboratory, if necessary. Record the situation in the Data Review Narrative, and note it for
       Regional Laboratory COR action.

    2.  Action regarding unsuitable blank results depends on the circumstances and origin of the blank.
       In instances where more than one blank is associated with a given sample, qualification should be
       based upon a comparison with the associated blank having the highest concentration of
       contaminant.
    3.  Some general "technical" review actions include:
       a.   For any blank (including Preparation Blanks and LEBs) reported with detects < CRQL, report
           detects < CRQL at the CRQL and qualify as a non-detect (U). For any blank (including
           Preparation Blanks and LEBs) reported with a detect <  CRQL, use professional judgment to
           qualify the sample results > CRQLs.  Non-detects should not be qualified.
       b.   For any blank (including Preparation Blanks and LEBs) reported with a negative result,
           < (-MDL) but > (-CRQL), carefully evaluate and determine its effect on the sample data. Use
           professional judgment to assess the data.
       c.   The blank analyses may not involve the same weights, volumes, or dilution factors as the
           associated samples. In particular, soil/sediment sample results reported on Form 1-IN will
           not be on the same basis (units, dilution) as the calibration blank data reported on Form 3-IN.
           It may be easier to work with the raw data and/or convert the ICB or CCB results to the same
           units as the soil/sediment samples for comparison purposes.
    4.  Specific "method" actions include:
       a.   If an ICB or a CCB result is > CRQL, the analysis should be terminated.  If the analysis was
           not terminated and the associated samples were not reanalyzed, non-detects should not be
           qualified.  Report detects < CRQL at the CRQL and qualify as non-detect (U). Report
           sample results that are > CRQL but < ICB/CCB Results at ICB/CCB Results and use
           professional judgment to qualify as non-detect (U) or unusable (R). Use professional
          judgment to qualify sample results that are > ICB/CCB Results.  Record the situation in the
           Data Review Narrative, and note it for Regional Laboratory COR action.
       b.   If an ICB or a CCB result is < (-CRQL), the analysis should be terminated.  If the analysis
           was not terminated and the associated samples were not reanalyzed, use professional
          judgment to qualify non-detects as estimated (UJ) or unusable (R). Use professional
          judgment to qualify detects < CRQL, or qualify as estimated low (J-). Use professional
          judgment to qualify sample results that are > CRQLs as estimated low (J-).


August 2014                                                                                  110

-------
Inorganic Data Review
Cyanide
       c.  If the concentration of any analyte in the Preparation Blank/LEB is > CRQL, the lowest
           concentration of that analyte in the associated samples must be > lOx the Preparation
           Blank/LEB concentration. Otherwise, all samples associated with that blank with
           concentrations < 1 Ox the Preparation Blank concentration and > CRQL should have been
           redistilled and reanalyzed. If the associated samples were not redistilled and reanalyzed,
           report the sample results at Preparation Blank Results; use professional judgment to qualify
           the results as estimated high (J+) or unusable (R). Report results < lOx the LEB
           concentration and > CRQL in the samples associated with the LEB at LEB Results; use
           professional judgment to qualify the results as estimated high (J+)  or unusable (R). Report
           detects < CRQLs in the  samples associated with the Preparation Blank/LEB  at CRQLs and
           qualify as non-detect (U). Non-detects and sample results that are  >  lOx the Preparation
           Blank/LEB Results should not be qualified. If the laboratory failed to redistill and reanalyze
           the samples associated with the Preparation Blank, record it in the  Data Review Narrative,
           and note it for Regional Laboratory COR action.
       d.  For any Preparation Blanks or LEBs reported with a negative result < (-CRQL), use
           professional judgment to qualify detects < CRQL, or qualify as estimated low (J-). Qualify
           sample results that are > CRQLs as estimated low (J-), and non-detects as estimated (UJ).
           Sample results that are > lOx CRQLs should not be qualified.

                          Table 31. Blank Actions for Cyanide Analysis
Blank
Type
ICB/CCB
ICB/CCB
ICB/CCB
ICB/CCB
Preparation
Blank/LEB
Blank Result
Detect < CRQL
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Detect < CRQL
Sample Result
Non-detect
Detect < CRQL
>CRQL
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but
< ICB/CCB Result
> ICB/CCB Result
Non-detect
Detect < CRQL
>CRQL
Non-detect
Detect < CRQL
>CRQL
Action
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
Use professional judgment
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at ICB/CCB Result and
qualify as non-detect (U) or unusable
(R)
Use professional judgment
Use professional judgment to qualify
as estimated (UJ) or unusable (R)
Use professional judgment or (J-)
Use professional judgment to qualify
as estimated low (J-)
No qualification
Report at CRQL and qualify as non-
detect (U)
Use professional judgment
August 2014
    111

-------
Inorganic Data Review
Cyanide
Blank
Type
Preparation
Blank/LEB
Preparation
Blank/LEB
Preparation
Blank/LEB
Blank Result
< (-MDL) but
> (-CRQL)
>CRQL
< (-CRQL)
Sample Result
Detect or non-detect
Non-detect
Detect < CRQL
> CRQL but
< 1 Ox the Preparation
Blank/LEB Result
> lOx the Preparation
Blank/LEB Result
Non-detect
Detect < CRQL
< lOx CRQL
> lOx CRQL
Action
Use professional judgment
No qualification
Report at CRQL and qualify as non-
detect (U)
Report at Preparation Blank/LEB
Result and use professional judgment
to qualify results as estimated high
(J+) or unusable (R)
No qualification
Qualify as estimated (UJ)
Use professional judgment or (J-)
Qualify results that are > CRQL as
estimated low (J-)
No qualification
August 2014
    112

-------
Inorganic Data Review                                                                   Cyanide

                               IV.    Duplicate Sample Analysis
A.  Review Items
    Cover Page, Form 6-IN, instrument printouts, and raw data.

B.  Objective
    The objective of duplicate sample analysis is to demonstrate acceptable method precision by the
    laboratory at the time of analysis.

C.  Criteria
    1 .   Samples identified as field blanks or Performance Evaluation (PE) samples cannot be used for
        duplicate sample analysis.
    2.   At least one duplicate sample shall be prepared and analyzed from each group of samples of a
        similar matrix type (e.g., water or soil) or for each SDG. Duplicates cannot be averaged for
        reporting on Form 1-IN. Additional duplicate sample analyses may be required by EPA Regional
        request.  Alternately, the Region may require that a specific sample be used for the duplicate
        sample analysis.
    3 .   A control limit of 20% for the Relative Percent Difference (RPD) shall be used for original and
        duplicate sample values > 5x the CRQL.
    4.   A control limit of the CRQL shall be used if either the sample or duplicate value is < 5x the
        CRQL.  The absolute value of the control limit  (CRQL)  shall be entered in the "Control Limit"
        column on Form 6-IN.  If both samples are non-detects, the RPD is not calculated for Form 6-IN.
    NOTE:     The above control limits are method requirements for duplicate samples, regardless of
               the sample matrix type. However, it should be noted that laboratory variability arising
               from the sub-sampling of non-homogenous soil  samples is a common occurrence.
               Therefore, for technical review purposes only, Regional policy or project Data Quality
               Objectives (DQOs) may allow the use of less restrictive criteria (e.g.,  35% RPD, 2x the
               CRQL) to be assessed against duplicate soil samples.
D.  Evaluation
    1 .   Verify, from the Cover  Page and the raw data, that the appropriate number of required duplicate
        samples were prepared  and analyzed for the SDG.
    2.   Verify, using Form 6-IN and the raw data, that the duplicate results fall within the established
        control limits.
    3.   Verify that a field blank or PE sample was not used for duplicate analysis.
    4.   Check the raw data and recalculate one or more of the RPD values using the following equation
        to verify that the results were correctly reported on Form 6-IN:

                                             IS-DI
        Where,
        RPD   =  Relative Percent Difference
        S      =  Sample result (original)
        D      =  Duplicate result
August 2014                                                                                113

-------
Inorganic Data Review                                                                    Cyanide

E.  Action
    NOTE:     For a duplicate sample analysis that does not meet the technical criteria, apply the action
               to all samples of the same matrix if the samples are considered sufficiently similar.
               Exercise professional judgment in determining sample similarity when making use of all
               available data, including: site and sampling documentation (e.g., location and type of
               sample, descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity,
               chlorine); and laboratory data for other parameters [e.g., Total Suspended Solids (TSS),
               Total Dissolved Solids (TDS), Total Organic Carbon (TOC), alkalinity or buffering
               capacity, reactive sulfide, anions].  Additionally, use the sample data (e.g., similar
               concentrations of analytes) in determining similarity between samples in the SDG.  The
               determinations are: 1) only some of the samples in the SDG are similar to the duplicate
               sample, and that only these samples should be qualified; or, 2) no samples are sufficiently
               similar to the sample used for the duplicate, and thus only the field sample used to
               prepare the duplicate sample should be qualified.
    1.  If the appropriate number of duplicate samples was not analyzed for each matrix using the correct
       frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the laboratory, if necessary.  Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action. Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If both original sample and duplicate sample results are >  5x the CRQL and the RPD is > 20%,
       qualify detects as estimated (J), and qualify non-detects as estimated (UJ).
    3.  If RPD > 100%, use professional judgment to determine if the associated  sample data should be
       qualified.
    4.  If both original sample and duplicate sample results are >  5x the CRQL and the RPD is < 20%,
       detects and non-detects should not be qualified.
    5.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate > CRQL,  qualify detects as estimated (J) and
       non-detects as estimated (UJ).
    6.  If the original sample or duplicate sample result is < 5x the CRQL (including non-detects) and the
       absolute difference between sample and duplicate < CRQL,  detects and non-detects should not be
       qualified.
    7.  If a field blank or PE sample was used for the duplicate sample analysis, note this for Regional
       Laboratory COR action. All of the other QC data must then be carefully checked.  Exercise
       professional judgment when evaluating the data.
    8.  Annotate the potential effects on the  data due to out-of-control duplicate sample results in the
       Data Review Narrative.
August 2014                                                                                 114

-------
Inorganic Data Review
Cyanide
                    Table 32. Duplicate Sample Actions for Cyanide Analysis
Criteria
Both original sample and duplicate sample results are > 5x
the CRQL and RPD > 20%*
RPD > 100%
Both original sample and duplicate sample results are > 5x
the CRQL and RPD is < 20%
Original sample or duplicate sample result < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate > CRQL*
Original sample or duplicate sample result < 5x the CRQL
(including non-detects) and absolute difference between
sample and duplicate < CRQL
Action
Detect
J
Use professional
judgment
No qualification
J
No qualification
Non-detect
UJ
Use professional
judgment
No qualification
UJ
No qualification
    *   The above control limits are method requirements for duplicate samples, regardless of the
       sample matrix type. However, it should be noted that laboratory variability arising from the sub-
       sampling of non-homogenous soil samples is a common occurrence. Therefore, for technical
       review purposes only, Regional policy or project DQOs may allow the use of less restrictive
       criteria (e.g., 35% RPD, 2xthe CRQL) to be assessed against duplicate soil samples.
August 2014
    115

-------
Inorganic Data Review                                                                    Cyanide

                                 V.     Spike Sample Analysis
A.  Review Items
    Cover Page, Form 5A-IN, Form 5B-IN, instrument printouts, and raw data.

B.  Objective
    The objective of the spiked sample analysis is to evaluate the effect of each sample matrix on the
    sample preparation procedures and the measurement methodology.

C.  Criteria
    1.   Samples identified as field blanks or PE samples cannot be used for spiked sample analysis.
    2.   At least one spiked sample (pre-distillation) shall be prepared and analyzed from each group of
        samples with a similar matrix type (e.g., water or soil), or for each SDG.

    3.   When the Matrix Spike recovery falls outside of the control limits and the sample result is < 4x
        the spike added, a post-distillation spike shall be performed.  An aliquot of the remaining
        unspiked sample shall be spiked at 2x the indigenous level or 2x the CRQL, whichever is greater.
    4.   The spike %R shall be within the established acceptance limits. However, spike recovery limits
        do not apply when the sample concentration is >4x the spike added. In such an event, the data
        shall be reported unflagged, even if the %R does not meet the acceptance criteria.
    5.   If the spiked sample analysis was performed on the same sample that was chosen for the duplicate
        sample analysis, spike calculations shall be performed using the results of the sample designated
        as the "original sample." The average of the duplicate results cannot be used for the purpose of
        determining %R.

    NOTE:     The final spike concentration required is presented in the method described in the SOW.
D.  Evaluation
    1.   Verify, using the Cover Page, Form 5 A-IN and raw data, that the appropriate number of required
        spiked samples was prepared and analyzed for the SDG.
    2.   Verify that a field blank or PE sample was not used for the spiked sample analysis.
    3.   Verify, using Form 5 A-IN and the raw data, that all pre-distillation  spiked sample results fall
        within the established control limits. If not, verify that a post-distillation spike was prepared and
        analyzed.
    4.   Recalculate, using the raw data, one or more of the %Rs using the following equation, and verify
        that the recalculated value agrees with the  laboratory-reported values on Forms  5A-IN & 5B-IN:

                                               SSR-SR
                                  %Recovery= ——	 x 100
                                                  oA.
        Where,
        SSR    =   Spiked Sample Result
        SR     =   Sample Result
        SA     =   Spike Added
    NOTE:     When the sample result is < MDL or reported as a non-detect, use SR = 0 only for the
               purpose of calculating the %R. The actual spiked sample results, sample results, and %R
               (positive or negative) shall still be reported on Forms 5A-IN & 5B-IN.
August 2014                                                                                 116

-------
Inorganic Data Review                                                                      Cyanide

E.  Action
    NOTE:    For a Matrix Spike that does not meet the technical criteria, apply the action to all
               samples of the same matrix if the samples are considered sufficiently similar. Exercise
               professional judgment in determining sample similarity when making use of all available
               data, including: site and sampling documentation (e.g., location and type of sample,
               descriptive data, soil classification); field test data (e.g., pH, Eh, conductivity, chlorine);
               and laboratory data for other parameters  (e.g., TSS, TDS, TOC, alkalinity or buffering
               capacity, reactive  sulfide, anions). Also use the sample data (e.g., similar concentrations
               of analytes) in determining similarity between samples in the SDG. The possible
               determination are: 1) only some of the samples in the SDG are similar to the Matrix
               Spike sample, and that only these samples should be qualified; or 2) no samples are
               sufficiently similar to the  sample  used for the Matrix Spike, and thus only the field
               sample used to prepare the Matrix Spike  sample  should be qualified.
    1.  If the appropriate number  of Matrix Spike samples was not analyzed for each matrix using the
       correct frequency, use professional judgment to determine if the associated sample data should be
       qualified; obtain additional information from the  laboratory, if necessary. Record the situation in
       the Data Review Narrative, and note it for Regional Laboratory COR action.  Detects should be
       qualified as estimated (J) and non-detects as estimated (UJ) if any of the frequency criteria is not
       met.
    2.  If a field blank or PE sample was used for the spiked sample analysis, note this for Regional
       Laboratory COR action. All of the other  QC data must then be carefully checked. Use
       professional judgment when evaluating the data.  Detects should be  qualified as estimated (J) and
       non-detects as estimated (UJ).
    3.  If the Matrix Spike recovery does not meet the evaluation criteria and a required post-distillation
       spike was not performed, note  this for Regional Laboratory COR action.
    4.  If the Matrix Spike %R is  < 30%, verify that a post-distillation spike was  analyzed (if required
       when sample concentration is < 4x spike added).  If the post-distillation spike %R is < 75% or the
       analysis was not performed, qualify detects as estimated  low (J-), and non-detects as unusable
       (R). If the post-distillation spike %R is > 75%, qualify detects as estimated (J) and non-detects as
       estimated (UJ).
    5.  If the Matrix Spike %R falls within the range of 30-74%, verify that a post-distillation spike was
       analyzed (if required when sample concentration  is < 4x  spike added).  If the post-distillation
       spike %R is < 75% or the  analysis was not performed, qualify detects as estimated low (J-), and
       non-detects as estimated (UJ).  If the post-distillation spike %R is > 75%, qualify detects as
       estimated (J), and non-detects as estimated (UJ).
    6.  If the Matrix Spike %R falls within the range of 75-125%, no post-distillation spike is required.
       Detects and non-detects should not be qualified.
    7.  If the Matrix Spike %R is  > 125%, verify that a post-distillation spike was analyzed (if required
       when sample concentration is < 4x spike added).  If the post-distillation spike %R is > 125% or
       the analysis was not performed, qualify detects as estimated high (J+); non-detects should not be
       qualified. If the post-distillation spike %R is < 125%, qualify detects as estimated (J); non-
       detects should not be qualified.
    8.  Annotate the potential effects on the data due to out-of-control spiked sample results in the Data
       Review Narrative.
August 2014                                                                                   117

-------
Inorganic Data Review
Cyanide
                      Table 33. Spike Sample Actions for Cyanide Analysis
Criteria
Matrix Spike %R < 30%
Post-distillation spike %R< 75%
Matrix Spike %R < 30%
Post-distillation spike %R> 75%
Matrix Spike %R 30-74%
Post-distillation spike %R< 75%
Matrix Spike %R 30-74%
Post-distillation spike %R> 75%
Matrix Spike %R > 125%
Post-distillation spike %R> 125%
Matrix Spike %R > 125%
Post-distillation spike %R< 125%
Matrix Spike %R < 30%
No post-distillation spike performed
Matrix Spike %R 30-74%
No post-distillation spike performed
Matrix Spike %R75-125%
Post-distillation not required
Matrix Spike %R > 125%
No post-distillation spike performed
Action
Detect
J-
J
J-
J
J+
J
J-
J-
No qualification
J+
Non-detect
R
UJ
UJ
UJ
No qualification
No qualification
R
UJ
No qualification
No qualification
    NOTE:    The above control limits are method requirements for spike samples, regardless of the
              sample matrix type.  However, it should be noted that laboratory variability arising from
              the sub-sampling of non-homogenous soil samples is a common occurrence. Therefore,
              for technical review purposes only, Regional policy or project DQOs may allow the use
              of less restrictive criteria (e.g., 10 %R and 150 %R for the lower and upper limits) to be
              assessed against spike soil samples.
August 2014
    118

-------
Inorganic Data Review                                                                   Cyanide

                    VI.     Regional Quality Assurance and Quality Control
A.  Review Items
    Form 1 -IN, instrument printouts, and raw data.

B.  Objective
    The objective is to use results from the analysis of Regional Quality Assurance/Quality Control
    (QA/QC) samples such as field blanks, PE samples, blind spikes, and blind blanks to determine the
    validity of the analytical results.
C.  Criteria
    Criteria are determined by each Region.
D.  Evaluation
    Evaluation procedures must follow the Region's Standard Operating Procedure (SOP) for data
    review. Each Region will handle the evaluation of PE samples on an individual basis. Compare
    results for PE samples to the acceptance criteria for the specific PE samples if possible.
    Calculate the RPD between field duplicates and provide this information in the Data Review
    Narrative.
E.  Action
    Any action must be  in accordance with Regional specifications and criteria for acceptable PE sample
    results. Note any unacceptable PE sample results for Regional Laboratory COR action.
August 2014                                                                                 119

-------
Inorganic Data Review                                                                     Cyanide

                               VII.    Overall Assessment of Data
A.  Review Items
    Entire data package, data review results, preparation logs, calibration standard logs, instrument logs,
    instrument printouts, and raw data (including any confirmation data).

B.  Objective
    The objective is to provide the overall assessment on data quality and usability.

C.  Criteria
    1.  Review all available materials to assess the overall quality of the data, keeping in mind the
       additive nature of analytical problems.
    2.  Reported analyte concentrations must be quantitated according to the appropriate analytical
       method, as listed in the method. All sample results must be within the linear calibration ranges
       per methods. Percent Solids (%Solids) must be properly used for all applicable matrix result
       calculations.
D.  Evaluation
    Examine the raw data to verify that the correct calculation of the sample results was reported by the
    laboratory. Distillation logs, instrument printouts, etc., should be compared to the reported sample
    results recorded on the appropriate Inorganic Summary Forms (Form 1-IN through Form 16-IN).
    1.  Evaluate any technical problems not previously addressed.
    2.  Examine the raw data for any anomalies (e.g., baseline shifts, negative absorbance, omissions,
       illegibility, etc.).
    3.  Verify that the appropriate methods and amounts were used to prepare samples for analysis. If
       reduced volumes were used, verify that the laboratory had received Regional Laboratory COR
       approval for the use of the reduced volume.
    4.  Verify that there were no transcription or reduction errors (e.g., dilutions, %Solids, sample
       weights, etc.) on one or more samples.  Recalculate %Solids for at least 10% of the samples and
       verify that the calculated %Solids agree with that reported by the laboratory.
    5.  Verify that the MDL is properly reported and it is not greater than the CRQL.

    6.  Verify that results fall within the calibrated range (Form 15-IN).
    7.  If appropriate information is available, assess the usability of the data to assist the data user in
       avoiding inappropriate use of the data.  Review all available information, including the Quality
       Assurance Project Plan (QAPP), focusing specifically on the acceptance or performance criteria,
       the SOPs, and communication with user concerning the intended use and desired quality of these
       data.
E.  Action
    1.  Use professional judgment to determine if there is any need to qualify data which were not
       qualified based on the QC criteria previously discussed.
    2.  Use professional judgment to qualify detects and non-detects if the MDL exceeds the CRQL.
    3.  If a sample is not diluted properly when sample results exceed the upper limit of the calibration
       range, qualify detects as estimated (J).
    4.  Write a brief Data Review Narrative to give the user an indication of the analytical limitations of
       the data.  Annotate any discrepancies between the data and the SDG Narrative for Regional
       Laboratory COR action. If sufficient information on the intended use and required quality of the
       data is available, include an assessment of the data usability  within the given context.
August 2014                                                                                   120

-------
Inorganic Data Review                                                                      Cyanide

    5.  If any discrepancies are found, notify the Regional Laboratory COR. The Regional Laboratory
       COR may contact the laboratory to obtain additional information for resolution. If a discrepancy
       remains unresolved, determine if qualification of the data is warranted.
August 2014                                                                                   121

-------
Inorganic Data Review	Cyanide

                                     VIII.   Calculations
Aqueous/Water Sample Concentration:
                                                         Vf
                            CN Concentration (ug/L) = C x -^ x DF

    Where,
    C      = Instrument response in |og/L CN from the calibration curve
    Vf     = Final prepared (absorbing solution) volume (mL)
    V      = Initial aliquot amount (mL)
    DF    = Dilution Factor

Soil/Sediment Sample Concentration:
                                                         Vf
               CN Concentration (mg/kg dry weight) = C x ——- x (1/1000) x DF
                                                        W x S

    Where,
    C      = Instrument response in |og/L CN from the calibration curve
    Vf     = Final prepared (absorbing solution) volume (mL)
    W     = Initial aliquot amount (g)
    S      = %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    = Dilution Factor

Adjusted MDL/Adjusted CRQL Calculation:
    To calculate the adjusted MDL or adjusted CRQL for aqueous/water samples, substitute the value of
    the MDL ((ig/L) or CRQL ((ig/L) into the  "C" term in the equation above.

    Calculate the adjusted MDL or adjusted CRQL for all soil/sediment as follows:
                                                              WM
                        Adjusted MDL or CRQL (mg/kg) = C x  —^- x DF


    Where,
    C      = MDL or CRQL (mg/kg)
    WM    = Minimum method required aliquot amount (1.00 g for Midi or 0.50 g for Micro)
    W     = Initial aliquot amount (g)
    S      = %Solids/100 (see Exhibit D - General Inorganic Analysis, Section 10.1.4)
    DF    = Dilution Factor
August 2014                                                                               122

-------
Inorganic Data Review                                                                Appendix A

                                  APPENDIX A: GLOSSARY

Analyte - The element or ion an analysis seeks to determine; the element of interest.
Analytical Sample - Any solution or media introduced into an instrument on which an analysis is
performed, excluding instrument calibration, Initial Calibration Verification (ICV), Initial Calibration
Blank (ICB), Continuing Calibration Verification (CCV), Continuing Calibration Blank (CCB), and
tunes. Note that the following are all defined as analytical samples: undiluted and diluted samples (EPA
and non-EPA); matrix spike samples; duplicate samples; serial dilution samples, analytical (post-
digestion/post-distillation) spike samples; Interference Check Samples (ICSs); Laboratory Control
Samples (LCSs); Performance Evaluation (PE) samples; and Preparation Blanks.
Analytical Services Branch (ASB) - The division of the United States Environmental Protection
Agency's (EPA) Office of Superfund Remediation and Technology Innovation (OSRTI) responsible for
the overall management of the Contract Laboratory Program (CLP).
Associated Samples - Any sample related to a particular Quality Control (QC) analysis.  For example,
for Initial Calibration Verification (ICV), all samples analyzed under the same calibration curve. For
duplicates, all Sample Delivery Group (SDG) samples digested/distilled of the same matrix.
Blank - An analytical sample that has negligible or unmeasurable amounts of a substance of interest.
The blank is designed to assess specific sources of contamination.  Types of blanks may include
calibration blanks, preparation blanks, and field blanks.  See the individual definitions for types of blanks.
Calibration - A set of operations that establish under specific conditions, the relationship between values
indicated by a measuring instrument and the corresponding known values. The calibration standards must
be prepared using the same type of reagents or concentration of acids as used in the sample preparation.
Calibration Blank - A blank solution containing all of the reagents, and in the same concentration, as
those used in the analytical sample preparation. This blank is not subjected to the preparation method for
Inductively Coupled Plasma-Atomic Emission Spectroscopy (ICP-AES) and Inductively Coupled
Plasma-Mass Spectroscopy (ICP-MS), but is digested for mercury and cyanide.  Calibration blanks are
used to verify that the instrument baseline is stable and the instrument is free of contamination.
Calibration Curve - A plot of instrument response versus concentration of standards.
Calibration Standards - A series of known standard solutions used by the analyst for calibration of the
instrument (i.e., preparation of the analytical curve).  The solutions may or may not be subjected to the
preparation method, but contain the same matrix (i.e., the same amount of reagents and/or preservatives)
as the sample preparations to be analyzed.
Case - A finite, usually predetermined number of samples collected over a given time period from a
particular site.  Case Numbers are assigned by the Sample Management Office  (SMO). A Case consists
of one or more Sample Delivery Groups (SDGs).
Continuing Calibration Blank (CCB) - A reagent water sample that is run 2 hours (ICP-AES, ICP-MS)
or every hour (Hg, CN) and designed to detect any carryover contamination.
Continuing Calibration Verification (CCV) - A  single parameter or multi-parameter standard solution
prepared by the analyst and used to verify the stability of the instrument calibration with time, and the
instrument performance during the analysis of samples.  The CCV can be one of the calibration standards.
However, all parameters being measured by the particular system must be represented in this standard and
the standard must have the same matrix (i.e., the same amount of reagents and/or preservatives) as the
samples.  The CCV should have a concentration in the middle of the calibration range and shall be
analyzed at the beginning of the day prior to the analysis of samples, and every 2 hours (1 hour for Hg
and CN).
August 2014                                                                                A-l

-------
Inorganic Data Review                                                                Appendix A

Contract Compliance Screening (CCS) - A screening of electronic and hardcopy data deliverables for
completeness and compliance with the contract.  This screening is performed under EPA direction by the
Contract Laboratory Program (CLP) Sample Management Office (SMO) contractor.
Contract Laboratory Program (CLP) - Supports the EPA's Superfund effort by providing a range of
state-of-the-art chemical analytical services of known and documented quality. This program is directed
by the Analytical Services Branch (ASB) of the Office of Superfund Remediation and Technical
Innovation (OSRTI) of the EPA.

Contract Required Quantitation  Limit (CRQL) - Minimum level of quantitation acceptable under the
contract Statement of Work (SOW).
Contractual Holding Time - The  maximum amount of time that the Contract Laboratory Program
(CLP) laboratory may hold the samples from the sample receipt date until analysis and still be in
compliance with the terms of the contract, as specified in the CLP Analytical Services Statement of Work
(SOW). These times are the same or less than technical holding times to allow for sample packaging and
shipping.
Duplicate - A second aliquot of a sample that is treated the same as the  original sample in order to
evaluate the precision.
Field Blank - Any sample that is submitted from the field and identified as a blank. A field blank is used
to check for cross-contamination during sample collection, sample shipment, and in the laboratory. A
field blank includes trip blanks, rinsate blanks, bottle blanks, equipment blanks, preservative blanks,
decontamination blanks, etc.

Field Duplicate - A duplicate sample generated in the field, not in the laboratory.
Initial Calibration - Analysis of analytical standards for a series of different specified concentrations;
used to define the quantitative response, linearity, and dynamic range of the instrument to target analytes.
Initial Calibration Blank (ICB) - The first blank standard analysis to confirm the calibration curve.
Initial Calibration Verification (ICV) - Solution(s) prepared from stock standard solutions, metals, or
salts obtained from a source  separate from that utilized to prepare the calibration standards. The ICV is
used to verify the concentration of the calibration standards and the adequacy of the instrument
calibration. The ICV should be traceable to National Institute of Standards and Technology (NIST) or
other certified standard sources when EPA ICV solutions are not available.
Interference  Check Sample (ICS) - A solution containing both interfering and analyte elements of
known concentration that can be used to verify background and interelement correction factors.
Internal Standard - A non-target  element added to a sample at a known concentration after preparation
but prior to analysis. Instrument responses to internal standards are monitored  as a means of assessing
overall instrument performance.
Laboratory Control Sample (LCS) - A matrix spiked at a known concentration. LCSs are analyzed
using the same sample preparation, reagents, and analytical methods employed for the EPA samples
received.
Matrix - The predominant material of which the sample to be analyzed  is composed. For the purposes of
this document, the matrices are aqueous/water, soil/sediment, and wipe.
Matrix Spike - Aliquot of a sample (aqueous/water or soil/sediment) fortified (spiked) with known
quantities of specific compounds and subjected to the entire analytical procedure to indicate the
appropriateness of the method for the matrix by measuring recovery.
Method Detection Limit (MDL) - The concentration of a target parameter that, when a sample is
processed through the complete method,  produces a signal with 99 percent probability that it is different
from the blank. For 7 replicates of the sample, the mean value must be 3.14s above the blank, where "s"
is the standard deviation of the 7 replicates.

August 2014                                                                                A-2

-------
Inorganic Data Review                                                                Appendix A

Percent Difference (%D) - As used in this document and the Statement of Work (SOW), is used to
compare two values.  It is the difference between the two values divided by one of the values multiplied
by 100.
Performance Evaluation (PE) Sample - A sample of known composition to the EPA; however,
unknown to the Contractor that is provided to evaluate Contractor performance.
Post-Digestion Spike/Post-Distillation Spike - The addition of a known amount of standard after
digestion or distillation (also identified as an analytical spike).
Preparation Blank - An analytical control that contains reagent water and reagents, which is carried
through the entire preparation and analytical procedure.
Regional Laboratory Contracting Officer Representative (Regional Laboratory COR) - The EPA
official who monitors assigned CLP laboratories (either inside or outside of the Regional Laboratory
COR's respective Region), responds to and identifies problems in laboratory operations, and participants
in on-site laboratory programs.
Relative Percent Difference (RPD) - As used in this  document and the Statement of Work (SOW) to
compare two values, the RPD is based on the mean of the two values, and is reported as an absolute value
(i.e., always expressed as a positive number or zero).
Relative Standard Deviation (RSD) - As used in this document and the Statement of Work (SOW), the
mean divided by the standard deviation, expressed as a percentage.
Sample - A single, discrete portion of material to be analyzed, which is contained in single or multiple
containers and identified by a unique Sample Number.
Sample Delivery Group (SDG) - A unit within a sample Case that is used to identify a group of samples
for delivery.  An  SDG is defined by the following, whichever is most frequent:

    •  Each 20 field samples [excluding Performance Evaluation (PE) samples] within a Case; or

    •  Each 7 calendar day period (3 calendar day period for 7-day turnaround) during which field
       samples in a Case are received (said period beginning with the receipt of the first sample in the
       SDG).

    •  All samples scheduled with the same level of deliverables.

    •  In addition, all samples and/or sample fractions assigned to an SDG must be scheduled under the
       same contractual turnaround time.  Preliminary Results have no impact on defining the SDG.
Samples may be assigned to SDGs by matrix (i.e., all soil/sediment samples in one SDG, all
aqueous/water samples in another) at the discretion of the laboratory. Laboratories shall take all
precautions to meet the 20 sample per SDG criteria.
Sample Management Office (SMO) - A contractor-operated facility operated under the SMO contract,
awarded and administered by the EPA. SMO provides necessary management, operations, and
administrative  support to the Contract Laboratory Program (CLP).
SDG Narrative - Portion of the data package which includes laboratory, contract, Case, Sample Number
identification, and descriptive documentation of any problems encountered in processing the samples,
along with corrective  action taken and problem resolution.
Serial Dilution - The dilution of a sample  by a factor of five. When corrected by the Dilution Factor
(DF), the diluted  sample must agree with the original undiluted sample within specified limits.  Serial
dilution may reflect the influence of interferents [Inductively Coupled Plasma (ICP) only].
Statement of Work (SOW) - A document which specifies how laboratories analyze samples under a
particular Contract Laboratory Program (CLP) analytical program.
August 2014                                                                                A-3

-------
Inorganic Data Review                                                               Appendix A

Technical Holding Time - The maximum amount of time that samples may be held from the collection
date until analysis.
Tune - A solution containing a range of isotope masses to establish ICP-MS accuracy, resolution, and
precision prior to calibration.  May also be called Instrument Performance Check sample (IPC).
August 2014                                                                                A-4

-------
Inorganic Data Review
                                                          Appendix B
CASE NO.
APPENDIX B: INORGANIC DATA REVIEW SUMMARY
                             SITE
LABORATORY.
MA NO.	
SOW NO.
                            NO. OF SAMPLES/MATRIX
                            SDG NO.	
                            REGION
REVIEWER NAME
                             COMPLETION DATE
REGIONAL LABORATORY COR
ACTION	
  REVIEW CRITERIA

1. Preservation and Holding
  Time
2. Tune Analysis
3. Calibration
4. Blanks
5. Interference Check Sample
6. Laboratory Control Sample
7. Duplicate Sample Analysis
8. Spike Sample Analysis
9. Serial Dilution
10. Internal Standards
11. Regional Quality Assurance
   and Quality Control
12. Overall Assessment of Data
                            FYI
                                  METHOD/ANAL YTE
              ICP-AES
ICP-MS
Mercury
Cyanide
August 2014
                                                                B-l

-------
Inorganic Data Review                                                                  Appendix B
                                This page is intentionally left blank.
August 2014                                                                                  B-2

-------