EPA-600/R-94/038d
    QUALITY ASSURANCE HANDBOOK
                    FOR
AIR POLLUTION MEASUREMENT SYSTEMS

    Volume IV — METEOROLOGICAL MEASUREMENTS
                as revised August 1989
           U. S. ENVMONMINTAL PROTECTION AGENCY
              Office of Research and Development
            Research Triangle F&k, North Carolina 27711

                                Wp Printed COT Recycled Paper

-------
                                                             Section No.   4. i
                                                             Revision No.    o
                                                             Date:   17 Sep 89
                                                             Page:    i  of  1
                                ACKNOWLEDGEMENTS
    This volume of the Qua 11 tv Assurance Handbook, has been prepared by Thoaas
J, Lockhart, CCM, of  the Meteorological Standards Institute on Fox Island,
Washington.  Sections 4.0.4.4, 4.1.8, 4.4.0, 4.5.0 and 4.6.0 were used from
EPA-60O/4-82-060 with only slight modification.  The contributions of the
authors of this work. Peter L. Flnkelsteln, Daniel A. Mazzarella, Thomas J.
Lockhart, William J.  King, and Joseph H. White are hereby gratefully
acknowledged.

     The first draft  of this volume received the contributions of the Critical
Review Panel who are:

                    Dr. C. Bruce Baker
                    Mr. Jack A. Bowen
                    Dr. Harold L. Crutcher, CCM
                    Prof. Gerald C. Gill, and
                    Mr. Alvin L. Morris.

    Work on this volume and the workshops which followed itspublication have
been immeasurably helped by the generous cooperation of the following companies
who made their products and services available to the author:

     AeroVironment Inc.            Campbell Scientific, Inc.
     Climatronics Corporation      Hollls International Limited
     Odessa Engineering            Rotronic Instrument Corp.
     Qualiroetrlcs, Inc.            Teledyne Geotech
     Waters Manufacturing, Inc.    R, M. Young Company

    This revised Volume  IV was produced under contract 68-02-4553, Mr. Jack A.
Bowen, Project Officer.
                                   DISCLAIMER

    Mention of trade  names or commercial  products does not constitute  EPA
endorsement or recommendation for use.

-------
                                         TECHNICAL REPORT DATA
                                    (Please read instructions on reverse before completing)
  1. REPORT NO.
   EPA-600/R-94-038d
                                                                             3. RECIPIENT'S ACCESSION NO.
  4, TITLE AND SUBTITLE
  Quality Assurance Handbook for Air Pollution Measurement Systems. - Volume
  IV: Meteorological Measurements
                    5. RETORT DATE
                    March 1995
                                                                            6. PERFORMING ORGANIZATION CODE
  7. AUTHOR(S)
  US EPA, Office of Research and Development, Atmospheric Research and
  Exposure Assessment Laboratory
                                                                             . PERFORMING ORGANIZATION REPORT NO.
  9. PERFORMING ORGANIZATION NAME AND ADDRESS
                                                                             10 PROGRAM ELEMENT NO.
   U.S. Environmental Protection Agency
   Office of Research and Development
   Atmospheric Research and Exposure Assessment Laboratory
   Research Triangle Park, NC 27711
                    !;. CONTRACT/GRANT NO-
                    68-02-4553
  12. SPONSORING AGENCY NAME AND ADDRESS
                                                                             IJ. TYPE OF REPORT AND PERIOD COVERED
   Director,
   Atmospheric Research and Exposure Assessment Laboratory
   U.S. Environmental Protection Agency
   Research Triangle Park, NC 27711	
                    14. SPONSORING AGBNCY CODE
                    EPA/2GO/Q4
  15. SUPPLEMENTARY NOTES
  16. ABSTRACT
  The Quality Assurance (QA) Handbook is comprised of five volumes: Volume 1 (Principles), Volume II (Ambient Air Methods),
  Volume III (Stationary Source Methods), Volume IV (Meteorological Measurements) and Volume V (Precipitation Measurement
  Systems). This volume of the QA Handbook recommends QA and QC procedures appropriate to meteorological measurement
  systems, including ground-based remote profilers such as Doppler Sodars, Doppler Radars, and Radio Acoustic Sounding Systems
  (RASS). This is likely to be the last version of this document published by the EPA Office of Research  and Development.
  17.
                                            KEY WORDS AND DOCUMENT ANALYSIS
                      DESCRIPTORS
                                                        b. IDENTIFIERS/OPEN ENDED TERMS
                                                                                                  c. COSATI Field/Group
  Meteorological Measurement
Aii" Pollution Measurement
  8. DISTRIBUTION STATEMENT

   Release Unlimited
19. SECURITY CLASS ^Report)
  Unclassified
                                                                                                  21. NO. OF PAGES
                                                        20. SECURITY CLASS (Pu
                                                           Unclassified
                                                                                                  22. PRICE
EPA Form 2220-! (Rev. 4-77)   PREVIOUS EDITION IS OBSOLETE

-------
                                           Section No. 1
                                           Date February 24, 1994
                                           Page 1


            OVERVIEW OP THE  INTERIM EDITION OF VOLUME IV

     The Quality Assurance  (QA) Handbook is comprised of five
volumes:  Volume I  (Principles), Volume II (Ambient Air Methods),
Volume III  (Stationary Source Methods), Volume IV (Meteorological
Measurements),  and Volume V (Precipitation Measurement Systems).
Much of the material in Volumes II, III and V are out-of-date and
some portions of these volumes have long been out-of-print.

     EPA is now preparing an updated version of the QA Handbook
series which will be available in September 1995.  To meet the
needs of the user community until the updated version is
available, EPA has published Interim Editions of Volumes I, II,
III, IV and V.   Each volume of the Interim Editions, is being
issued as a complete unit with out-of-date sections either
deleted or modified using addendum sheets and handwritten
notations in the text.

     This volume and the other four volumes of the Interim
Edition of the QA Handbook are available at no charge from:

               USEPA/ORD
               Center for Environmental Research Information
               26 West Martin Luther King Drive
               Cincinnati, Ohio  45268

     Since this volume was updated in 1989, this interim edition
is being published without changes.  The updated version,  which
will be available in September 1995, will contain information on
-eteorological systems for measuring upper air meteorological
conditions.

     The user of the QA Handbook is cautioned to bear in mind
"That the information provided in the handbook is for guidance
purposes only.   EPA regulations are published in the Code of
Federal Regulations  (CFR),  When information in the CFR conflicts
with information in the QA Handbook, the CFR shall be considered
~he authoritative and legally bonding document.
                              William J. Mitchell
                                   Chief
                         Quality Assurance Support Branch

-------
                                                             Section No.  4.ii
                                                             Revision No.    0
                                                              Date:   17 Sep 89
                                                              Page:    1  of  2
                                   VOLUME IV
                               TABLE OF CONTENTS
Section
4.i    ACKNOWLEDGEMENTS
41U   TABLE OF CONTENTS

4.0    GENERAL PROGRAM REQUIREMENTS
       FOR METEOROLOGICAL MEASUREMENTS
    4.0.0   OUTLINE,  PURPOSE AND OVERVIEW
    4.0.1   GLOSSARY
    4.0.2   STATE OF THE ART
    4.0.3   DATA REQUIREMENTS
    4.0.4   MEASUREMENT REQUIREMENTS
    4.0.5   REFERENCES
4.1
   GENERAL ASPECTS OF QUALITY ASSURANCE
   FOR METEOROLOGICAL MEASUREMENTS
      1.0   OUTLINE AND SUMMARY
      1.1   PLANNING FOR A QUALITY ASSURANCE PROGRAM
      1.2   ORGANIZATION OF AUTHORITY AND
            RESPONSIBILITY
      1.3   QUALITY CONTROL GUIDELINES
      1.4   TRACEABILITY PROTOCOL
      1.5   ESTIMATING PRECISION AND ACCURACY
      1.6   SYSTEM AUDITS
      1.7   PERFORMANCE AUDITS
      1.8   DATA VALIDATION PROTOCOL
      1.9   QA REPORTS AND CORRECTIVE ACTION
    4.1.10  REFERENCES
4.2    QA FOR WIND SPEED, WIND DIRECTION AND
       TURBULENCE
       .3
       .4
4.2.0
4.2.1
4.2.2
4.2.
4.2.
4.2.5
4.2.6
4.2.7
4.2.8
4.2.9
OUTLINE AND SUMMARY
TYPES OF INSTRUMENTS
SPECIFICATIONS
ACCEPTANCE TESTING
INSTALLATION
CALIBRATION
OPERATIONS, MAINTENANCE AND QC
PERFORMANCE AUDIT METHODS
ESTIMATING ACCURACY AND PRECISION
REFERENCES
4.3    QA FOR TEMPERATURE AND TEMPERATURE
       GRADIENTS (AT)
    4.3.0   OUTLINE AND SUMMARY
    4.3.1   TYPES OF INSTRUMENTS50
    4.3.2   SPECIFICATIONS
    4.3.3   ACCEPTANCE TESTING
    4.3.4   INSTALLATION
    4.3.5   CALIBRATION
    4.3.6   OPERATIONS, MAINTENANCE AND QC
                                                       Pages  Rev.  Date
                                                 1
                                                 2

                                                25

                                                 S
                                                 2
                                                 6
                                                 3
                                                 7
                                                 2

                                                33

                                                 3
                                                 3
                                                 1
 4
 6
 2
 1
 7
 2
 2

83

 4
 4
32
 5
 S
 4
 2
20
 1
 3
                                                         3
                                                         4
                                                         4
                                                         2
                                                         1
                                                         9
                                                         3
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89

-------
                                                             Section No.   4.ii
                                                             Revision No.     0
                                                              Date:    17 Sep 89
                                                              Page:     2  of  2
Section
                                   VOLUME IV
                               TABLE OF CONTENTS
    4.3,7   PERFORMANCE AUDIT METHODS
    4.3.8   ESTIMATING ACCURACY AND PRECISION
    4.3.9   REFERENCES
4.4.0   QA FOR PRECIPITATION MEASUREMENTS

4.5.0   QA FOR RELATIVE HUMIDITY OR DEW POINT
        TEMPERATURE
   j
4.6.0   QA FOR SOLAR RADIATION MEASUREMENTS

4.7.0   QA FOR ATMOSPHERIC PRESSURE
Pages  Rev.  Date
  8
  1
  1
  7

 10
  8

  2
0
0
0
0
9/17/89
9/17/89
9/17/89
9/17/89

9/17/89
0   9/17/89

0   9/17/89

-------
                                                            Section No.  4.0.0
                                                            Revision No.     0
                                                            Date:    17 Sep 89
                                                            Page:     1  of  5
                                 Section 4.0
                       GENERAL PROGRAM REQUIREMENTS FOR
                         METEOROLOGICAL MEASUREMENTS
                                   OUTLINE
Section
4.0.0  OUTLINE,  PURPOSE AND OVERVIEW
       OF  THE QUALITY ASSURANCE HANDBOOK

4.0.1  GLOSSARY

4.0.2  STATE OF  THE ART

       4.0.2.1   AUDITOR SURVEY
       4.0.2.2   INTERVIEW SUMMARY

4.0.3  DATA REQUIREMENTS

       4.0.3.1   REGULATORY PROGRAMS
       4.0.3.2   RESEARCH PROGRAMS
       4.0.3.3   CONTINGENCY PROGRAMS

4.0.4  MEASUREMENT REQUIREMENTS

       4.0.4.1   MEASUREMENT SYSTEM
       4.0.4.2   DOCUMENTATION
       4.0.4.3   MOUNTING
       4.0.4.4   SITING

4.0.5  REFERENCES
Pages Rev.   Date


5     0     9/89

2     0     9/89

6     0     9/89
            9/89
            9/89
      0
9/89
               PROTECTED UNDER INTERNATIONAL COPYRIGHT
               ALL RIGHTS RESERVED
               NATIONAL TECHNICAL INFORMATION SERVICE
               U.S. DEPARTMENT OF COMMERCE

-------
                                                             Section No.  4.0.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  5


                      PURPOSE AND OVERVIEW OF THE QUALITY
                              ASSURANCE HANDBOOK

    The purpose of this volume of the QA Handbook is to provide information and
guidance for both the meteorologist and the non-meteorologist who must make
Judgments about the validity of data and accuracy of measurement systems.   Care
has been taken to provide definitions to help those making these judgments to
coMnunicate without ambiguity.  Methods are described in the handbook which
will objectively define the quality of measurements so the non- meteorologist
can, communicate with the meteorologist or environmental scientist or engineer
with precision of meaning.

    The first section of the handbook contains a special glossary of terns
necessary to meteorology and quality assurance.  Following that is an analysis
of the state of the art from information and interviews of those practicing QA
in the air quality field.  The final parts of the first section define some of
the requirements for gathering data which a QA effort can compare to the
practice of acquiring data.

    The second section is devoted to quality assurance and quality control as
it is applied to meteorological problems.  This section is somewhat Independent
of the variable being Measured.   Where the variable is important it Is treated
individually.

    The final six sections are variable-specific.   The nost Important wind
Measurement is covered In considerable detail.  The temperature measurement
section concentrates on the temperature difference measurement used for
stability determination. " The final four sections cover to an adequate depth
the measurement of humidity, radiation, precipitation and surface air pressure.
Examples are given where possible to help explain the methods and problems to
be found in programs of collecting meteorological data and assessing data
validity.

    The need for common understanding is critical for the practice of quality
control (QC) and quality assurance (QA).  This is achieved in part by the
definitions of the language used within the discipline.  From that vocabulary,
the details of the systems and procedures are defined in terms of the necessary
goals.

    There are a variety of QA/QC definitions in the literature and in common
usage.  Volume I, Section No. 1.3 and Appendix A provide some general
definitions.  Section 1.4 shows how the elements of QA are distributed and
wnere in the section they are described.  The well known "quality assurance
wheel" is shown in Figure 1.4.1.  The following discussion of definitions Is
broader based to include meteorological requirements and explicit between QA
and QC.

    The structure shown in Figure 4.0.0.1 below is from ANSI/ASQC Q9Q-19S7;
American National Standard, Quality Management and Quality Assurance Standards
Guidelines for Selection and Use.  The definitions in the glossary (4.0.1) and
the following descriptions are structured to fit Figure 4.0.0.1 and the
practices of meteorological measurement.

-------
                                                             Section No,  4.0.0
                                                             Revision No,     0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  5

    [Organizational
    \  Structure /
    /Confidence\
    1   To The   I
    ^Management/
                      QUALITY MANAGEMENT ASPECTS*
       —?-»• QUALITY, §YSTEM  X
                  QUALITY
                 CONTROL
                 ASPECTS"
  1	"	INTERNAL
   \   "     QUALITY
   \  *, ASSURANCE
          ASPECTS
    \    •  —-^r •;;v^;,r.^W
     \   "'• S*''<   ^  ^i^l^S
       N^x_ /r*;4^"^>«^^S
                                                           (Policy)
                                                         / Operational \
                                                         I  Techniques I
                                                         \And Activities/
   (When Required
    By A Contract)
 EXTERNAL QUALITY
ASSURANCE ASPECTS

  CONRDENCE TO
    PURCHASER
Figure 4.0.0.1
    Structure Of Quality (r*prSnt«d with per«t*»ion
fro« the American Society Tor Quality Control)
    This figure will be described as It applies to a meteorological company
which manufacturers instruments and provides a variety of services.  It can
also apply to government organizations and temporary project-oriented
cooperative organizations.  At the top of the figure and the top of any
organized effort toward quality are the QUALITY MANAGEMENT ASPECTS.  These
aspects comprise the policy statement for the organization expressed by
management.  The statement is in writing as a company policy and signed by the
president of the company so that there will be no misunderstanding or confusion
about the quality goals of instruments so that they do produce valid data.  A
performance audit, then, is a challenge both to the instrument and to the
operator to independently verify that the measurement system is "in control."
Just as with system audits, the auditor is primarily a teacher and trainer.
The audit method should be the best possible method.  The operator should be
encouraged to upgrade the calibration methods to do a better job.

    Often the operator has no experience with meteorological instruments.
Often they are well qualified instrument technicians, but the auditor is an
expert,  or should be.  A mutually valuable goal is for the operator to learn
what methods are necessary and most complete and adopt those for the
calibration procedure.  When the instruments are all working perfectly it is
because they are getting the experienced attention it takes for "in control"
operation.   At this point, the audit becomes a spot checking operation
producing documentation from an independent individual verifying this "in
control" operation.

    If some basic ground rules are followed, the audit is maximized as a
learning exercise.  One rule is that the operator does all the instrument
handling.   There is a general reluctance to handle unfamiliar instruments.
They might get broken or changed in some mysterious way.  The way  to become

-------
                                                              Section No.  4,0.0
                                                              Revision No.     0
                                                              Date;    17 Sep 89
                                                              Page:     4  of  5


familiar with  them  is  to work with them.   The  safest  environment in which to
gain this experience is in the presence of someone  who  is  familiar  with the
instruments.

    Good audit methods are as close to laboratory tests as a field  site and the
ingenuity of the auditor will allow.   This inevitably requires a tower mounted
instrument  to  be taken down,  complete with cable or substitute cable, so that
tests can be made in a physically convenient environment.   For wind sensors,
the bearing condition  is of vital importance.   This can be measured only when
the sensor  can be connected to the torque instrument  with  complete  control and
freedom to  move.  It Is not a proper test to try on a tower or mast.
Performance audits, in exactly the same way as calibrations,  challenge parts of
tbe system.  Ideally,  one wants to challenge all of the system, but that is
often impossible.   Known free atmospheres are  not available from NBS.
Controlled  atmospheres like a wind tunnel or a thermal  chamber or a "sun" lamp
only challenge a part  of the system.   They leave out  or drastically change the
important coupling  function.   Even controlled  atmosphere devices, such as a
wind tunnel, are not available to the auditor  in the  field.   All one can do is
iapose a known condition such as the rate of rotation for  an anemometer, and
measure the system  response.   This type of audit assumes that the
•anufacturer* s generic transfer function applies to this sensor, or assume that
earlier wind tunnel tests of this sensor still apply  (a good assumption if the
sensor is not  damaged).

    Another method  is  the ASTM collocated transfer  standard method.  This is
the most complete method from the standpoint of total system error  sources but
it has two  drawbacks.   First, it Is limited to the  conditions that  prevail
during the  audit.   Secondly,  it is very sensitive  to  exposure or siting bias.
 It requires careful guidelines pointing out potential bias sources  and ways to
watch for them in the  data.  These are covered in  the variable-specific
sections.

    A performance audit program using experienced  independent auditors, whether
 internal or external  to the organization, is  the first  step toward  establishing
_a quality plan if one  does not already exist.   The  goal of the measurement
program is  to  have  documented data.  The performance  audit will point out areas
required to get  the system "In control."  The  auditor can  help implement the
establishment  of a  quality system, or its key  elements, in order to achieve the
necessary on-going  activities to keep the measurement system "in control"
continuously.

    The survey which  led off the work of revising  this  handbook exposed some
confusion in  the community of meteorological  auditors with regard to the
difference  between  performance audits and calibrations. A form letter was
ccaposed to discuss these differences and to ask for  new numbers of audits
conducted.  The  letter used the definitions found  in  the glossary  (4.0.1) and
expanded on them with  examples.  The principal difference  is independence of
responsibility.  Some  organizations perceived  the  documentation of  the
rendition of  the system "as found" as a performance audit  and the adjustment of
 the system  to  acceptable operating conditions, documented  "as left," as a
 calibration.   Thus, a  single individual could  both audit and calibrate during
 tie same visit.  By any accepted standard of  quality  systems definition, this
whole process  of testing and adjustment is a calibration.   This properly
documented  calibration is the basis for claims of  data  validity.  All the

-------
                                                             Section No. 4,0.0
                                                             Revision No,    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  5

performance audit adds is an independent assurance that the calibrations were
thoroughly done and that the documents are complete and accurate.  Such
assurance must be entirely free of potential influence.

    The letter described situations where a single company can be structured to
provide both calibration and auditing services, but cautioned that the
independence of the auditor requires a management structure insulating the
auditor from the budgetary concerns of the operating organization.  Responses
to this letter were few and in no case was the distinction challenged.     All
agreed with the concept of independence for QA audits.  Regulators should
acknowledge the distinction and require true independence.

    Situations arise where the fundamental principle of independence between
calibration and audit services is difficult to follow.  Small agencies may not
be able to contract for independent audits.  In the interest of documented
validity of data for all parties, innovative arrangements among different
agencies should be promoted.  The individual who operates and calibrates
instruments for Agency A might be asked to audit the Agency B instruments in
exchange for the operator at Agency B auditing the Agency A instruments.  This
practice would have the further benefit of stimulating communication about and
standardization of good audit methods.

-------
                                                             Section No. 4.0,1
                                                             Revision No.    0
                                                             Date:   IT Sep 89
                                                             Page:    1  of  2
4.0.1       GLOSSARY FOB METEOROLOGY AND QA/QC
ACCURACY - is the degree of agreement of a measurement (or an average of
measurements of the same thing), X.wlth an accepted reference or true value, T,
usually expressed as the difference between the two values, X - T, or the
difference as a percentage of the reference or true value, 1QO(X-T)/T, or
sometimes expressed as a ratio, X/T.

CALIBRATION - is a MEASURE of conformance to or discrepancy from a
specification or set of criteria for an instrument or system if necessary and
an ADJUSTMENT of the instrument or system to conform to the specification or
criteria.   A calibration may be performed by a person or agency within the
operating organization.

DAMPING RATIO (i\) - The damping ratio is calculated from the overshoot ratio
(Q).  121
                              lnf l
DELAY DISTANCE CD) - The distance the air flows past a wind vane during the
 "te it takes the vane to return to 50 percent of the Initial displacement.  (2]

EXTERNAL QUALITY ASSURANCE - is the activity designed to provide the purchaser
with confidence in the quality of what is being purchased.

INTERNAL QUALITY ASSURANCE - is the activity designed to provide management
with confidence that the quality system is operating and the management policy
is being carried out.

INVERSION (+AT3 - is the inverted lapse rate or an increase of air
temperature with height.  There is no general limit for inversion strength.

LAPSE RATE (-AT) - is the normal decrease of air temperature with height
limited by the auto convection rate of  3,4 C/100 m.
OVERSHOOT (9) - The ratio of the amplitude of two successive deflections of  a
wind vane as it oscillates about the equilibrium position after release from an
offset position of ten degrees, as expressed by the equation
                          &
                           n

where 9  and 6,     are the amplitudes of the n and n+1 deflections,
       n      (n+1)
respectively.

-------
                                                             Section No, 4.0.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2

PERFORMANCE AUDIT - is a report of conformance to or discrepancy from a
specification or set of criteria determined by  a person or agency separate
froai and independent of the operating organization.

PRECISION -  is the standard deviation of a series of measured values, X,»

about the mean measured value, X. (see 4.1.5,13

QUALITY ASSURANCE - All those planned and systematic actions necessary to
provide adequate confidence that a product or service will satisfy given
requirements for quality,  [ii

QUALITY CONTROL - The operational techniques and activities that are used  to
fulfill requirements for quality, til

QUALITY MANAGEMENT - That aspect of the overall management function that
determines and implements the quality policy, [il

QUALITY POLICY - The overall quality intentions and direction of an
organization as regards quality, as formally expressed by top management.  til

QUALITY SYSTEM - The organizational structure, responsibilities, procedures,
processes, and resources for implementing quality management, [t]

REPRESENTATIVENESS - is the extent to which a set of measurements taken in a
space-time domain reflects the actual conditions in the same or different
space-time domain taken on a scale appropriate for a specific application. (41

STARTING THRESHOLD (S , m/s) - The lowest speed at which a vane will turn  to
within 5  of 8  (the true direction) from an initial displacement of 10°.  taj
              B

STARTING THRESHOLD (U , m/s) - The lowest speed at which a rotating anemometer

starts and continues to turn and produce a measurable signal when mounted  in
its normal position.  [31

Hi  ANSI/ASQC, 1987a.-  Quality Management and Quality Assurance
    Standards - Guidelines for Selection and Use. ANSI/ASQC Q90-1987. American
    Society for Quality Control, Milwaukee, HI 53203.

121  ASTM, 1985b:  Standard Test Method for DETERMINING THE DYNAMIC PERFORMANCE
    OF A WIND VANE. (Draft 8 of D22.ll) Amer. Soc. for  Testing and Materials,
    Philadelphia, PA 19103.

E3J  ASTM, 1985a:  Standard Test Method for DETERMINING THE PERFORMANCE OF  A CUP
    ANEMOMETER OR PROPELLER ANEMOMETER. (Draft 6 of D22.ll) Amer. Soc. for
    Testing and Materials, Philadelphia, PA 19103.

-------
                                                             Section No. 4,0.2
                                                             Revision    No. 0
                                                             Date;   17 Sep 89
                                                             Page;    1  of  6
4.0.2  STATE OF THE ART
    The achievement of predicted quality for a product or service can be the
delegated responsibility of an identifiable part of an organization.  The
practice of elevating quality to a management staff level is relatively new.
Th« value of quality control and the umbrella management structure of quality
assurance became clear when products* purchased against a specification, were
rejected by the purchaser.  When the cost of rework or scrap absorbs the
profit,  an alternative will be found.  The alternative is to do it right the
first time and the path to that goal involves training, in-process inspection,
final Inspection and all of the other QA functions designed to minimize scrap
and rework.

    The QA profession grew during World War II and thereafter as the U.S.
Government became a significant purchaser using comprehensivespecifications,
like the well known Mil-Specs.  In the " 60sand '70s the practice of planned
obsolescence and using the customer as the final inspector set up our
industries for failure against foreign competition with higher quality
standards.  The successful foreign producers, using the quality principles
developed in the United States, caused a resurgence of quality awareness.

    The Environmental Protection Agency (EPA) recognized the need to set
standards, develop standard methods and materials, and produce a system of
quality assurance to support validity claims for the data being collected in
response to the Clean Air Act.  In 1976 a Quality Assurance Handbook for Air
Pollution Measurement Systems: Volume I. Principles was published (EPA, 1976).
la 1977, Volume II. Ambient Air Specific Methods {EPA, 1977a> and Volume III,
Stationary Source Specific Methods (EPA, 1977b) were published.  This program
addressed the Criteria Pollutants which were covered by federal law.
Meteorological measurements were recognized as supportive to the Criteria
Pallutant measurement program but they were secondary.

    When the Clean Air Act was amended, permission for growth of source
strength (and thereby growth of Industry) was granted as a con- sequence of
diffusion model predictions based on input meteorological data.  Now, the law
recognized the requirement for valid and represent- ative meteorological data
am the need for a structure to provide documented assurance of validity.  In
1983, Volume IV. Meteorological Measurements (EPA, 1983) was added to the
Handbook family.

    Almost all of the QA work provided by the private sector was geared to air
aad source chemistry.  When meteorology was added to the technical requirements
list, a variety of solutions were applied by a variety of individuals with a
variety of technical backgrounds.  The original Volume IV. was like a
background guidebook for taking meteorological measurements and general
suggestions for how QA and QC might be applied to the requirement for valid
data.  This revision of Volume IV. is intended to be more specific and more
informative and more in the spirit of the other three volumes.  It has not, and
cannot as yet, specify standard methods.  A greater success with predictive
medals is necessary before knowledge will exist which can dictate the standard
tee-hods to assure valid input data.  The premise of this Volume IV. is that
measurements worth talcing are worth taking right.

-------
                                                             Section No. 4,0.2
                                                             Revision    No. 0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  6
       4.0.2.1 Auditor Survey

       This handbook is intended to document the methods currently in use in
•eteorological QA/QC and to point to methods which are optimum for meeting the
requirements suggested or defined in various EPA publications.  A starting
point toward this goal is a survey of all those active in performance and
system auditing of meteorological measurement programs.  Figure 4.0.2.1 is a
copy of the survey form sent to as many people with experience In auditing as
could be found.  The initial list, shown by company and location in Table
4.0.2.1, grew considerably with help from all the EPA Regions and many state
and local agencies.  The number of survey forms returned from each company is
also shown.

                   Table 4.0.2.1  - Original  Survey List
                  COMPANY

      * AeroVironment,  Inc.
        Dames & Moore
        Desert Research Institute
      * Enviro. Monitoring & Services Inc.
        Environmental Research & Tech.
      * Environmental Research & Tech.
        Galson Technical Services, Inc.
        Meteorological Standards Inst.
        Research and Evaluation Assoc.
      * Research Triangle Institute
        Roy F. Weston,  Inc.
        RTP Associates
        Technical Environmental Enter.
        Tennessee Valley Authority
      * TRC Environmental Consultants
  CITY/STATE
Monrovia, CA
Atlanta, GA
Reno. NV
Thousand Oaks, CA
Fort Collins, CO
Concord, MA
E. Syracuse, NY
Fox Island, WA
Chapel Hill, NC
RTP, NC
West Chester, PA
Denver, CO
Aurora, CO
Muscle Shoals, AL
E. Hartford, CT
NUMBER

   4
   2
   2
   1
   1
   4
   1
   1

   1
   1
   1
   1
   1
           indicates companies chosen for in-depth interview
        Of the 70 or so forms sent originally or copied and distributed within
an organization, 49 forms were returned.  The summary of these responses is
shown numerically on Figure 4.0.2.1.  The number of audits represented by the
survey is 12,195, where the definition of an audit is the challenge of one
instrument measuring a meteorological variable.  Each respondent was asked to
qualify himself by specialty, using three or more if necessary but Indicating a
priority of 1,2 or 3.  Some managers reported for their organization of
auditors.  The responses to the questions were not weighted by numbers of
audits.  As with most surveys, a few points are useful but action should not be
based on the survey results.  Of the 49 survey forms returned with data, 21
came from the original list, 9 came from local, state or federal agencies and
19 came from others.  Of this 19, 5 came from utilities in the Northeast U.S.
(3 from Pennsylvania Power & Light) showing a close relationship to Regional
Meteorologists and interest in QA/QC in the area.

-------
                                                        Section No,  4.0.2
                                                        Revision    No.  0
                                                        Date;   17 Sep 89
                                                        Page:    3  of  6
                 Survey  of Meteorological Measurement QC/QA  People
                         (please print  or type your responses)
      NAME	COMPANY.
      ADDRESS	         ADDRESS
      PHONE   (     >                             <      )
      I am * „	meteorologist, _ chemist,  	environmental  scientist,  	QA/QC
      professional, __instrument technician,   electronic* tech.,  __engineer,
        wet.tech., 	modeler. 	manager.   da€a  analyst,    field hand,  (use
      172,3 if you are- more than one)   this • information will  be summarized
      without the us* of names or  companies or agencies so please  be candid.
         Consider an audit or challenge as a QA  observation of  the response
      of an instrument to a known  input and consider a calibration as an
      OPERATIONAL testing and adjustment, as necessary, of an instrument.:
      1. If one meteorological audit  is defined  as a challenge  to  one
         variable or one variable  of  a system, about how  many audits have you
         performed in 1980-1984 	,  1985	,  1986	, 1987	?
      2. Did you 3*4 usual Iv _2_sometimes  «{«fcnever use  a **ritten procedure?
      3. If a calibration is defined  as the testing and adjustment of one
         variable or one variable  in  a system, how many calibrations have you
         performed in 198O-19B4 	,  1985	,  1986  	, 1987 	?
      4. Did vou|~.:9_ywuallv |,Jflt.*ofl|etimes  _JL_never use  the manufacturer's
         calibration procedure?
      5. When you perform an audit, do you require the operator to remove the
         sensors froai their mounted position?  _tl.Yes  £?QMQ |.5£.Sometimes
      6. Do you require the operator  to re—connect the sensor to the system
         when it is presented for  audit? SfiYes f^Mo
      7. If 6. is yesf is the re-connection made with jj*'2*ne operational
         cable, _Q;.a substitute cable or  JtLeither?
      8. Do you gffusual 1 v j_|_sometimes _/<3Lnever measure  the starting torque
         of each anemometer bearing assembly and transducer?
      9. Do you (^ usual 1 y /jQsometimes /^.never measure  the starting torque
         of each wind vane bearing assembly and  transducer?
      1O. Do you _£_u»ually /^.somet i mes ff*^never use  the collocated transfer
         standard method for auditing a Mind instrument?
      11. Do you 3£J_ usual 1 y / j^some t i mes  Jjj*_never find the audited instrument
         meets the required specification?
      12. Do you challenge anemometers with known rates of rotation?
         /jjjjfr No — If yes, how many  speeds  	_?  „__Synchronous or
      13. Do you challenge direction  vanes with  a dividing wheel?
         |j*l No - If yes, how many  angles	?  	CW»	CCW,	both.
      14. Will you fill out a more detailed questionaire  as a contribution to
         the quality of this project? fjfeoYes  _t_No
Figure 4.0.2.1  Survey Form and Response Summary

-------
                                                             Section No. 4.0.2
                                                             Revision    No. 0
                                                             Date:   17 Sep 89
                                                             Pagei    4  of  6

        The number of audits, sorted by the technical background rated by the
respondent as #1, is shown in Table 4.0.2.2.  It is comforting to note that the
largest number of auditors consider themselves meteorologists first.  While the
largest number of audits were reported by persons considering themselves
Managers first, it is likely that those they managed were distributed like the
rest of the group.  One organization reported by a manager listed 3,600 audits.
It is likely that the discipline of the person doing most of the performance
audits is meteorologist.

       The information from each question gives some feeling about how the
audits reported were conducted.  Question 2 shows that 76% of the auditors back
their work with a written procedure.  Question 4 shows that 64% of the auditors
usually use the manufacturer's calibration procedure.  This answer pertains to
the calibration function which most auditors perform as a separate part of
their job.

                         Table 4.0.2.2 - Survey Summary
Meteorological Performance Audits
#1 Specialty Number
Meteorologist 11
Engineer 8
QA/QC 7
Instrument Tech. 5
Manager 4
Environ. Sci. 4
Electronics Tech. 4
Chemist 3
Data Analyst 1
No indication 2
Total 49
Average number per year
Percent change
1980-84
1,291
253
387
129
2,115
224
510
352
0
0
5,261
1,052
	
1985
647
194
193
702
543
70
165
220
0
0
2,102
2,102
100
1986
473
188
237
102
551
130
181
256
0
30
2,256
2,256
7
1987
623
206
318
60
644
91
212
192
0
30
2,576
3,435
52
Total
3,034
841
1,135
669
3,853
515
1,068
1,020
0
60
12, 195


       In question 5, 43% of the auditors either do not physically inspect the
sensor or do so by performing the operator's function of climbing the tower and
raaoving the sensor.  Volume IV. should reduce that percentage to zero.
Question 6 suggests that most auditors (67%) do both a physical and an
operational challenge of the sensor when it Is down from the tower.  The
conditional question 7 shows a preference for the operational cable (71%) over
a substitute cable.

       Questions 8 and 9 show only 43% of the auditors usually measure the
smarting torque of the anemometer and only 41% usually measure the direction
vane starting torque.  It looks like when an auditor decides to make this
measurement, both sensors are included.  Several respondents answered "never"
but indicated that they were getting equipment to make the measurement in the
future.   Other auditors inspect the bearing assemblies with educated fingers
welch tell the auditor whether or not they are "all right" but fail to provide
numerical or objective documentation.

-------
                                                             Section No. 4.0.2
                                                             Revision    No. 0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  6
       Question 10 shows that only 19% rely on the collocated transfer standard
(CIS)  method for auditing wind instruments.  Volume IV. should help increase
that number.  Question 11 may mean that 70% of the instruments audited are
working within specification, or it may mean that the audit methods used are
not rigorous enough to find the discrepancies.  Ths fact that half the audits
do not include a torque measurement,  the only method short of a wind tunnel to
challenge starting threshold, points to the latter possibility.  Volume IV.
should help to improve audit methods toward a standard practice so that this
question,  asked in the future, will provide an unambiguous answer.

       Questions 12 and 13 show a difference in challenging speed and
direction.   There were 73% of the auditors who indicated the use of a simulated
speed  to challenge an anemometer.  The most common number of speeds was two
(52X), followed by three (21%)t then one (15%), and finally four or more (12%).
There  were 86% who indicated a synchronous motor was used.  This near unanimity
is probably because of the availability of synchronous motors and the lack of
availability of simple measurement systems.  The measured method Is the only
choice where good commercial power is not available.

       The direction challenges were not as uniform.  Of the 23 who indicated
the number of angles used, seven said 4, five said 6, four said 8, three said
5, and one each said 1, 12, 16 and 18.  There is no consensus there.  All but
two said they used both clockwise and counterclockwise rotation.  The two used
clockwise.

       If the survey did one thing, it demonstrated the need for guidance
toward an acceptable standard of performance auditing.  It also demonstrated a
recognition of need to move toward that goal and a willingness to help in the
process.   Only one of 47 said no to question 14,

       4.0.2.2 Interview Sumsary

       After the survey results were in, a series of visits was planned to talk
to private sector organizations which had a recognized role in quality
assurance of meteorological measurements.  The first organization visited was
AeroVironroent, Inc. of Monrovia, California.  The half-day discussion with four
AV auditors was a frank exchange of methods currently in use, shortcomings of
Volume IV and suggestions for the content of the revised Volume IV.  The
principle of starting torque measurements of anemometer and wind vane shafts as
a field substitute for starting threshold wind speed determination in a wind
tunael was accepted.  The principle of operators doing all the climbing or
handling of sensors was currently practiced.

       The second interview was at Environmental Monitoring and Systems, Inc.
in Thousand Oaks, California.  Half-day discussions with two meteorological
auditors reinforced the belief that some organizations were advanced in the
practice of meteorological QA.  Comprehensively written audit procedures were
followed.   Questions of the difference between an audit and a calibration were
correctly answered with authority.  The need for uniform expectations or
requirements was expressed in the context of competitive bidding for providing
audit  services.  It was felt that the new Volume IV could help buyers of

-------
                                                             Section No. 4.0,2
                                                             Revision    No, 0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  6

services to specify a scope of work in enough detail to both assure a
comprehensive service and provide a fair bidding competition.

       The third interview was with Environmental Research and Technology {now
±NSR) in Concord,  Massachusetts, perhaps the largest of the five organizations
In terms of meteorological services and auditing.  Some different concepts and
practices were found, particularly in the area of starting threshold
determination.  Seven meteorologists, field auditors and QA specialists were
present during the half -day discussion.  This organization was a leader in the
field of providing meteorological monitoring services to industry.  As a result
of the history of providing all services including design, Installation,
operation, data summarization and QA auditing, an interesting discussion was
lield on the subject of independence between operators/calibrators and auditors.

       The fourth interview was with the head of the field operations
department of TRC Environmental Consultants in E. Hartford, Connecticut.  This
organization was also a leader in providing full meteorological monitoring
services.   Their procedures developed in a different way.  They calibrated
their sensors by wind tunnel testing in their calibration facility and employed
a regular replacement of sensors in the field.  All of the performance auditing
related to sensors was done by QA personnel in the calibration facility.  This
aethod requires a spare set of sensors be available for each client.  The
nethods described in Volume IV for calibrating or auditing in the field are not
necessary if you have a wind tunnel and employ the Interchangeable sensor
aethod.

       The final interview was with a meteorologist/auditor from the Research
Triangle Institute of Research Triangle Park, North Carolina.  Since two of the
original Volume IV authors were at RTI when the work was written, it was not
surprising to find the methods employed to be acceptable standard methods.  The
level of quality of the field standards used In auditing was the highest, as it
    with most of these organizations.
       This series of interviews provided valuable insights and confirmations
about the best methods to use for meteorological quality assurance practices.
It showed the field to be well practiced at the level of the largest and best
consulting organizations.  The task for Volume IV Is to provide a basis for a
standard practice in this field at all levels, and to provide a measure by
mich those practicing in the field can be judged by those with the final
authority to accept or reject data on the basis of documented validity.

-------
                                                             Section No, 4.0.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  3
4.0.3  DATA REQUIREMENTS
    There are a variety of reasons why meteorological data are collected.  Some
reasons relate to regulatory requirements or national monitoring programs.
Some data are collected for the purpose of research.  Some data are collected
against the contingency that they may be needed at some future time.  Sometimes
data are collected for one reason and then used for other reasons.

    The philosophy upon which this volume rests is the belief that data need to
have an estimation of uncertainty before the numbers can be dignified by the
title "data.*  The estimation might be a simple declaration such as "The
meteorological measurement program was operated in conformance with PSD
guidelines."  This cites the accuracy requirements for PSD as the uncertainty
level for the data and promises that the documentation required for validity
claims for a PSD application will be available to back these data.  When such
an estimation exists and rests on documentation of performance, decisions can
be made as to whether or not these data are appropriate for the application.

    4.0.3.1 Regulatory Programs

            4.0.3.1.1  PSD

            The regulatory program used in this document, and to some extent in
the On-Site Meteorological Program Guidance for Regulatory Modeling
Applications (EPA, 1987b)» is the Ambient Monitoring Guidelines for Prevention
of Significant Deterioration (PSD) (EPA, 1987a).  This is the most explicit
guideline and one requiring a quality of instrument performance available only
from sensitive instruments.  Recently it seems to be used for other programs as
a "standard" of specification.

            Most air quality dispersion models described in the Guideline on Air
Quality Models (EPA, 1986) used for PSD applications are Gaussian models
requiring input data which represent the conditions at the site of interest and
which follow a prescribed data content and format.  The models require five
meteorological inputs.  They are:

            1) Wind speed - representing the average wind speed at 10 n above
the ground (and additional heights for elevated sources) during each hour
calculated by a scalar average or mean of samples taken during the hour,
usually in 15 minute increments.  The samples may be the integrated wind run
during the sample period (one or two seconds is often used) or instantaneous
samples of speed.  A resultant vector magnitude does not represent the initial
dilution for which the Gaussian model uses wind speed.

            2) Wind direction - representing the average wind direction at 10 m
above the ground (and additional heights for elevated sources) during each hour
calculated by carefully averaging samples of wind direction or by calculating
the resultant vector direction using unity as the wind speed for each sample.
A resultant vector direction does not represent the distribution of direction
samples which occurred during the hour.

            3) Temperature - representing the air temperature at the standard 2
m height above ground (and additional heights for elevated sources).

-------
                                                             Section No. 4,0.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  3

            4) Stability class - representing the site of interest can be
estimated by a variety of schemes.  Turner (1970) describes a method based on
the observation of cloud cover, ceiling height and wind speed along with a
known solar elevation angle.  Estimations may also be based on the distribution
of wind direction (slgma theta) or on the vertical temperature gradient (delta
T).  Current research Is investigating whether or not the Turner method
stability class can be estimated with measurements of solar radiation (daytime
sky cover substitute) and 2 to 10 a delta temperature (nighttime sky cover
substitute) along with wind speed, latitude and date. The method which will be
acceptable for the site of interest is determined by the regulatory authority.

            5) Mixing height - may be estimated by a method described by
Holzworth (1972).

            The PSD guideline provides accuracy and performance requirements
for wind speed, wind direction, temperature, vertical temperature difference,
and solar radiation along with humidity, precipitation and visibility.

            Measurements for PSD permitting may, In some cases, be continued
after the new source begins to verify the estimations made by modeling.
Continued monitoring requires the same QA/QC efforts as the permit phase
required.

            4.0.3.1.2  Other Programs
            Meteorological measurements may be made to augment air quality
neasurements made to demonstrate compliance with the National Ambient Air
Quality Standards (NAAQS) or to monitor trends.

       4.0.3.2  Research Programs


       Meteorological data networks may be installed for special model
validation studies.  The same kind of QA/QC efforts are necessary for these
programs but they are usually applied on a shorter time scale since the
programs are relatively short in duration and the need for documented accuracy
could not be greater.

       Data representativeness is a critical question as the terrain increases
in complexity.  Research looks into the number and location of measurement
sites and the applicability of certain types of instruments to characterize the
eoaplex  (turbulent or stratified) air flow systems.  Different instruments,
such as Doppler sodars for remote vertical sounding, sonic anemometers for
saall eddy size sensitivity and low threshold speeds, laser anemometers for
long path length integration, and even the old standby bivane, are examined to
try and optimize the detection of important aspects of flow measurement for
model Inputs or verification.

    Meteorological data are used to find correlations with aerometric
measurements in a continuing search for better forecasting capability.

       4.0.3.3  Contingency Programs

       Industry may choose to monitor meteorological variables at
representative sites on their property to document the local air flow

-------
                                                             Section No.  4.0.3
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  3

conditions in case excessive concentrations are observed which might
erroneously be attributed to their source.   While such programs may not fall
under any regulatory requirements, the use of the data for its contingency
purpose  requires documentation sufficient to verify the data accuracy.

      Other programs may be exploratory to see how on-site data compare with
public data from other sites (airports or state or local agency stations).
Such questions of representativeness cannot be convincingly answered if the
on-site  data does not come from suitably sensitive instruments, properly
calibrated and maintained and subject to QA/QC effort designed to document data
validity.

       It is possible to select, install, operate and document on-site
measurement systems to meet PSD requirements.  Public data from airports may
differ from valid on-site data for three reasons.   Representativeness deals
with different meso-scale structures in the surface layer flow driven,  in part,
by the larger synoptic flow.  It is common to find that airport measurements do
not represent other sites just a few miles away because the flow is different.
It is also common to find airport data to be different from on-site data
because  the airport data is essentially an instantaneous sample (a one minute
average) taken within ten minutes of the end of the hour while the on-site data
for the  same hour includes samples from the entire hour.  Finally,-the airport
instruments are selected to serve aviation where low wind speeds are of no
             Airport instruments do not meet PSD requirements.

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  7

4.0.4  MEASUREMENT REQUIREMENTS

       4.0.4.1  Measurement System

                4.0.4.1.1  Sampling

                The usual period of time assigned a data value is an hour.  It
is getting common to use a shorter intermediate period in the process of
generating the hour value.  Fifteen minutes is the recommended intermediate
period.   The fifteen minute values are usually calculated from samples taken
during the period.  The number of samples is related to the accuracy with which
the samples represent the true value for the period.  It has been found that
when the mean is estimated by 60 samples, the sampling error is 5 to 10
percent.   Also, when the standard deviation is estimated by 360 samples, the
sampling error is also 5 to 10 percent.  For this reason, the required number
of samples for sigma theta, the standard deviation of the wind direction is
equal or greater than 360 (EPA, 1987b).


                If a fifteen minute period is used and if 360 samples are
required within that period, a simple calculation shows the maximum time
between samples is 2.5 seconds.  How a sample is taken and what it represents
is also a consideration.  If a continuous output voltage Is available, as with
wind direction, a sample of the voltage can be taken at any time.  If the wind
speed is calculated by measuring rate of rotation by counting pulses during a
fixed time, as is common for systems with the sensor directly connected to a
data logger (without a dedicated signal conditioner), the "sample" Is really
the average for the fixed time.  If samples are taken once a. second and the
anemometer provides three pulses per revolution and the anemometer turns one
revolution for every 0.3 meters of air that goes through it, each pulse will
represent 0.9 m/s.  If samples are taken every 2 seconds, the resolution of the
wind speed sample becomes 0.45 m/s.  A 15 minute period at 2 second sampling
will have 450 samples.  The average wind speed will be accurate with a
resolution of better than 0.1 m/s.   The variance of the wind speed samples may
be influenced by the 0.45 m/sresolution of the sample.

                Quality assurance considerations should incilude the
determination and documentation of the sampling procedures used in generating
the reported hourly data values.

                4.0.4.1.2  On-Line Processing

                There are two on-line processing programs commonly used in air
quality meteorology.  One is the program used to combine wind speed and
direction samples for an hour.  The other is the program used to calculate or
estimate sigma theta.

                The QA role is to determine what these programs do and judge
the suitability of the programs for the measurement application.   The field of
software QA for meteorology is in it infancy and methods are not standardized
as yet.

                4.0.4.1.3  Data Handling

                There is a need to provide data in certain formats for some
applications.   If the data are machine processable in the final measurement

-------
                                                             Section No.  4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  7

step,  any reconfiguration required will be handled by a program which can be
subject  to software QA.

                If any hand entry work is required, a data handling QC step is
required to be sure that errors of transcription do not enter the data base.

      4.O.4.2  Documentation

      If there is a requirement to show evidence of data validity, the process
of documenting the various QA» QC, and operational activities is important.
The added time such documentation takes is usually proportional to the degree
of preparation and training which has been applied,

                4.0.4.2.1  Station Log

                The station log is the journal of all happenings at the
aeasurenent site.  These include visits where no problems are found, scheduled
calibration visits and findings, unscheduled maintenance tests and repairs, and
audits.   It is a truism that there are never enough field notes to reconstruct
with certainty what happened in the past.  Planning for the day when such a
reconstruction may be necessary can save a long period of data from being
discarded because of inadequate documentation.

                4.0.4.2.2  Reports

                Any activity effecting the measurement system should be
reported.  This procedure allows responsible individuals to follow these
activities without visiting the measurement site or witnessing calibrations and
audits.   It also provides input to a file of activities related to the system.
Reports  should include calibrations, audits, discrepancies found and corrected,
modifications or upgrades and the like.  Reports do not need to be exhaustive
or glossy but they do need to be as factual and succinct.


      4.0.4.3  Siting and Mounting

                4.0.4.3.1  Introduction

                Although good instrumentation is a necessity, proper site
selection is critical to obtain good meteorological data.  It is, from an
absolute error point of view, much more important than proper placement of any
other kind of air monitoring equipment.  Poor placement can and has caused
errors of 180  in wind direction, and can cause major errors in any other
meteorological variable, including wind speed, temperature, humidity, and solar
radiation.

                The purpose of this section is to offer guidance  in assessing
the suitability of meteorological monitoring sites.  The guidance given  is
based principally on standards set by the World Meteorological Organization
(«MD, 1971), the Federal Meteorological Handbook No. 1 (NWS, 1979) and the
Tennessee Valley Authority (TVA,  1977).  For an understanding of  flow around
obstacles and their potential bias to wind data, see Hosker  (1984).

                Proper siting is part of the total quality control program.  Of
ccurse,  as in many other monitoring activities, the Ideal may not be attainable
aid, in  many urban areas where air quality studies are traditionally done,  it
will be  impossible to find sites  that meet all of the siting criteria.   In

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date;    17 Sep 89
                                                             Page:     3  of  7


those cases, compromises must be made.  The important thing to realize is that
the data will be compromised, but not necessarily in a random way.   It is
incumbent upon the agency gathering the data to describe carefully the
deficiencies in the site and, if possible, quantify or at least evaluate the
probable consequences to the data.

                4.0.4.3.2  Instrument Siting

                The primary objective of instrument siting is to place the
instrument in a location where it can make precise measurements that are
representative of the general state of the atmosphere in that area, consistent
with the objectives of the data collection program.  Because most atmospheric
properties change dramatically with height and surroundings, certain somewhat
arbitrary conventions must be observed so that measurements can be compared.
In this section, conventions published by the World Meteorological Organization
0*O, 1971) have been adopted wherever possible.  Secondary considerations such
as accessibility and security must be taken into account, but should not be
allowed to compromise data quality.
                     4.0.4.3.2.1  Wind Speed and Direction

                     "The standard exposure of wind instruments over level,
open terrain is 10 m above the ground" (WHO, 1971), however optimum measurement
height may vary according to data needs.  Open terrain is defined as an area
where the horizontal distance between the instrument and any obstruction is at
least ten (10) times the height of that obstruction.  An obstruction may be
man-made (such as a building) or natural (such as a tree)(Figure 4.0.3.1).
                                tor
Figure 4.0.4.1
Siting wind instruments; a 10 m tower located at  'east
10 times the height of obstructions away from those
obstructions (not to scale),
                     The wind instrument should be securely mounted on a mast
that will not twist, rotate, or sway.   If it is necessary to mount the wind
instrument on a roof of a building,  it  should be mounted high enough to be out
of the area in which the air flow is disturbed by the building.  This is
usually 1.5 times the height of the  building above the roof so  that It is out

-------
                                                             Section No. 4.0,4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  7

of the wake of the obstruction.  This  is not a good practice, however, and
should only be resorted  to when absolutely necessary.  Sensor height and its
height above the obstructions, as well as the character of nearby obstructions,
should be documented.

                     4.0.4.3.2.2  Temperature and Humidity

                     Temperature and humidity sensors should be mounted over a
plot of open level ground at  least 9 meters in diameter.  The ground surface
should be covered with non-irrigated or unwatered short grass or, in areas
where grass does not grow, natural earth.  The surface must not be  concrete or
asphalt or oil soaked.   The standard height for cllmatologlcal  purposes is 1.25
to 2 m, but different heights may frequently be required  in air quality
studies.

                     The sensors should not be closer to  obstructions such as
trees and/or buildings than a distance equal to four times their height.  They
should be at least 30 m  from  large paved areas and not close to steep slopes,
ridges, or hollows.  Areas of standing water should also  be avoided.  Louvered
instrument shelters should be orien.ted with the door opening toward true north,
in the northern hemisphere.

                     4.0.4.3.2.3  Radiation

                     Solar and whole sky radiation measurements should be taken
in a  location free from any  obstruction to the measurements.   This means there
should be nothing above  the horizontal plane of the sensing element that would
cast a shadow on it.  Neither should the instrument be near light colored walls
or artificial sources of radiation.  Usually a tall platform or roof of a
building is  the most suitable location.

                     4.0.4.3.2.4  Precipitation

                     A rain gage should be mounted on level ground  so that the
mouth or opening is horizontal.  The gage should be shielded from the wind but
not placed in an area where  there will be excessive turbulence  caused by the
shield.  For example, a  good  location  would be an opening in an orchard or
grove of trees where the wind speed near the ground is reduced  due  to the
canopy effect, but a location that  is  mostly open except  for one or two trees
would not be good because  of  the strong eddies that could be set up by  the
trees.  This admittedly  requires a good deal of subjective judgment but it
cannot be avoided.  Obstructions to the wind should not be closer than  two to
four times the obstruction height from the instrument.   In open areas,  a wind
shield such  as that used by  the U.S. National Weather Service should be used.
The ground surface around  the rain gage may be natural vegetation or gravel.
It should not be paved,  as  this may cause splashing into  the gage.  The gage
should be mounted a minimum of 30 cm above the ground and should be high enough
so that it will not be covered by snow.

                     4.0.4.3.2.5  Meteorological Towers

                      It  is frequently  necessary to measure some meteorological
variables at more than one height.  For continuous measurements or  where  the
height requirement  is not  too restrictive, towers may offer the most
advantageous measurement platform.

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   1? Sep 89
                                                             Page:    5  of  7

                     Towers should be located in an open level area (see Table
4.0.4,1)  representative of the area under study.  In terrain with significant
topographic features, different levels of the tower may be under the influence
of different meteorological regimes at the same time.   Such conditions should
be well documented.

      Table 4.0.4.1   I.Unit* on Terrain and Obstacles Near Towers
Distance Slope
from tower (between)
(m) (%)
0-15 ±2
15- 30 ±3
30-100 ± 7
100-300 ± 11

Source: TVA, 1977
Max, obstruction or
vegitation height
(m)
0.3
0.5-1.0 (most veg.<0.3)
3.0
10 x ht. must be less
than distance
to obstruction
                     Towers should be of the open grid type of construction,
such as is typical of nost television and radio broadcast towers.  Enclosed
towers,  stacks, water storage tanks, grain elevators, cooling towers, and
 'tilar structures should not be used (Mollo-Christensen, 1979).  Towers must
   rugged enough so that they may be safely climbed to install and service the
instruments.   Folding or collapsible towers that make the instruments available
to be serviced or calibrated at the ground are desirable provided they are
sufficiently j-igj^ to hold the instruments in the proper orientation and
attitude during normal weather conditions.

                     Wind instruments should be mounted above the top of the
tc«er or on booms projecting horizontally out from the tower.  If a boom is
used,  it should support the sensor at a distance equal to twice the maximum
disaeter or diagonal of the tower away from the nearest point on the tower.
The boom should project into the direction which provides the least distortion
for the most important wind direction.  For example, a boom mounted to the east
of the tower will provide least distortion for north or south winds.  One may
wisa to consider having two sets of instruments at each level, located on
opposite sides of the tower.  A simple automatic switch can choose which set of
daza to use (NASA, 1968).  Documentation of the tower should include the
orientation of the booas.
                     Temperature sensors must be mounted on booms to hold them
away from the tower,  but a boom length equal to the diameter of the tower is
Efficient.   Temperature sensors should have downward facing aspirated shields.
Tbe booms aust be strong enough so that they will not sway or vibrate
excessively in strong winds.  The best vertical location on the tower for the
secsors  is at a point with a minimum number of diagonal cross members, and away
frr« major horizontal cross members.  Even with these precautions, data
-~tained while the wind blows from the sector transected by the tower may not
   free  from error,
                     These instrument siting suggestions may seem to preclude
tbe use  of many air monitoring sites that otherwise would be desirable, but

-------
                                                             Section No. 4,0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  7

this need not be the case.  In siting air quality monitors that are to be used
for long—term trend analysis or large geographic area coverage, it may be
perfectly acceptable to have some or all of the meteorological equipment at a
different location that better meets the large-scale requirements of the study.
As long as both sites are in the same area of interest and meet their
respective siting criteria, this should present no problems.  When the air
quality data are to be used for short-term diffusion model validation or
studies of short-term levels from specific sources, however, a meteorological
station should be located in the vicinity of the air quality sensor.

                4.0.4.3.3  Station Siting

                Besides care in selecting the local environment of a
•eteorological sensor, it is important that care be taken in selecting station
location with respect to major man-made and topographic features  such as
cities, mountains, large bodies of water, etc.  Meteorological variables are
obviously affected by the large-scale surrounding features.  The  effect of
cities has been studied extensively (Ito, 1972; Vukovich, 1971; U.S.PHS, 1961).
Documented effects include a decrease in an average wind speed, decrease in
atmospheric stability, increase in turbulence, increase in temperature, and
changes in precipitation patterns.  These changes will obviously  have an effect
on the evaluation and interpretation of meteorological and air quality data
taken in an urban area.

                Even more pronounced are the effects of large natural features
(Slade, 19631.  Besides their obvious effect on humidity, oceans  and large
bodies of water are usually at a different temperature than the nearby  land.
This generates the well known land- and sea-breezes which, in many coastal
areas, dominate the wind patterns.  There are also often simultaneous
differences in cloud cover between the oceans and nearby land surfaces.  This
difference in thermal lag, insolation, and changes in surface roughness and
vertical temperature structure can have a profound effect on atmospheric
stability (SethuRaman, 1974).

                The effects of mountains and valleys on meteorological
variables and atmospheric dispersion continue to be studied.  Two of the more
interesting recent papers are by Kao (1974) and Hunt  (1978).  Well-known
effects include the channeling of flow up or down a valley, the creation of
drainage flow, the establishment of lee-waves, and an increase in mechanically
generated turbulence.  All of these effects and others can play a major role in
the transport and dispersion of pollutants.

                The important point is that almost any physical object has an
effect upon atmospheric motion.  In fact, it is probably impossible to  find a
site that is completely free from obstruction.  This being the case, it  is the
responsibility of the person choosing a monitoring site to have in mind  the
various forces at work and to choose a site that will be most representative of
the area of interest.  If the area is a valley or a sea coast, then the
meteorological instruments should be in that valley or near the coast; not on a
nearby hilltop or inland  30 km at a more convenient airport site.  Of course
one must also keep in mind the vertical structure of  the atmosphere.  Minds
measured at the bottom of a 100 m valley will not represent the winds at the
top of a 200 m stack whose base happens to be 5n that valley.

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    7  of  7


                The choice of a station for meteorological data collection must
be made with a complete understanding of the large-scale geographic area, the
sources being investigated, and the potential uses of the data.  Then rational,
Informed choices can be made.

                Once they are made, the site should be completely documented.
This should include both small- and large-scale site descriptions, local and
topographic maps (1:24,000 scale), photographs of the site (if possible), and a
written description of the area that is adequately represented by this site.
This last point is most important for it will allow a more rational
interpretation of the data.  It might state, for example, that a site
adequately represents a certain section of a particular valley, the suburban
part of a given city, or several rural counties.  Whatever It is, the nature of
the site should be clearly described in a way that will clear to those who will
be using the data in the future.

-------
                                                             Section No. 4.0.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Eage:    1  of  2
4.0.5     REFERENCES
EPA,  1976:  Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol. I, Principles, EPA-600/9-76-005.  Office of Research and
    Development, Res. Triangle Park, MC 27711.

EPA,  1977a:  Quality Assurance Handbook for Air Pollution. Measurement
    Systems, Vol. II, Ambient Air Specific Methods, EPA~60Q/4-77-027a.   Office
    of Research and Development, Res. Triangle Park, NC 27711.

EPA,  1977l>:  Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol. Ill, Stationary Source Specific Methods,
    EPA-600/4-77-027b.  Office of Research and Development,
    Res.  Triangle Park, NC 27711.

EPA,  1982:  APTI Course SI:  409, Basic Air Pollution Meteorology, Student
    Guidebook EPA 450/2-82-009.  Air Pollution Training Institute,
    Environmental Research Center, Research Triangle Park, NC 27711.

EPA,  1983:  Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol. IV, Meteorological Measurements, EPA-600/4-82-Q60.  Office
    of Research and Development, Res. Triangle Park, NC 27711.

EPA,  1986:  Guideline on Air Quality Models (Revised).  OAQPS, U.S.
    Environmental Protection Agency, Research Triangle Park, NC.,
    EPA-450/2-78-027R (NTIS PB 288-783).

EPA,  1987a:  Ambient Monitoring Guidelines for Prevention of
    Significant Deterioration  (PSD), EPA-450/4-87-007, Office of Air
    Air Quality Planning and Standards, Res. Triangle Park, NC 27711

EPA,  1987b:  On-Site Meteorological Program Guidance for Regulatory
    Modeiing Applications, EPA-450/4-S7-013, Office of Air
    Quality Planning and Standards, Res. Triangle Park, NC 27711.

Holzworth, G. C., 1972:  Mixing Heights, Wind Speeds, and Potential for
    Urban Air Pollution Throughout the Contiguous United States.  Office of
    Air Programs, U.S. Department of Health, Education and  Welfare, Research
    Triangle Park, NC.  Publication Nol. AP-101.

Hosker, Jr. ,R. P.,1984:  Flow and Diffusion Near Obstacles, Chapter 7.
    Atmospheric Science and Power Production,   Darryl Randerson,
    Ed. DOE/TIC-27601, pp 241-326.

Hunt,  J.  C. R., W. H. Snyder and R. E. Lawson, Jr., 1978;   Flow structure
    and turbulent diffusion around a three dimensional hill:  part 1.
    EPA-6QO/4-78-041. U.S. Environmental Protection Agency.

I to,  Masashi et al. , 1972:  An examination of local wind measurements in
    cities.  Annual Report of the Tokyo Metropolitan District Public Damage
    Research Institute,3(3), pp. 27-32 (APTIC 41974 TR 186-73).

-------
                                                             Section No. 4.0.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2

Kao, S.  K. ,  H. N. Lee and K. I. Smidy, 1974:  A preliminary analysis of the
    effect of mountain-valley terrains on turbulence and diffusion.
    Symposium on Atmospheric Diffusion and Air Pollution, American
    Meteorological Society, Santa Barbara, CA, pp. 59-63.

Mollo-Christensen, E.> 1979:  Upwind distortion due to probe support in
    boundary layer observations.  Journal of Applied Meteorology, 18(3),
    pp.  367-370.

NASA, 1968:   Meteorological Measuring and Recording Equipment
    Description, Calibration and Maintenance Procedures for NASA's 150
    Meter Meteorological Tower Facility, Kennedy Space Center, Florida,
    National Aeronautics and Space Administration, GP-465.

NWS, 1977:  Federal Meteorological Handbook No. 1, Surface Observations.
    U.S. Department of Commerce, Washington, D.C,

SethuRaman,  S., R. M. Brown, and J. Tlchler, 1974:  Spectra of
    atmospheric turbulence over the sea during stably stratified
    conditions.  Symposium on Atmospheric Diffusion and Air Pollution,
    American Meteorological Society, Santa Barbara, CA, pp. 71-76.

Slade, D. H,, 1968:  Meteorology and Atomic Energy.  U.S. Atomic Energy
    Commission, TID-24190.

Turner,  D. B. ,  1970:  Workbook of Atmospheric Dispersion Estimates,
    Revised.  Office of Air Programs, U.S. Department of Health, Education
    and Welfare, Research Triangle Park, NC.  Publication No. AP-26.

TVA, 1977:  Criteria for meteorological measurement site acceptance
    and/or preparation (unpublished).  Tennessee Valley Authority.
    U.S.Public  Health Service, 1961:  Air Over Cities. A symposium,
    Cincinnati, Ohio.

Vukovich, F. M. ,  1971:  Theoretical analysis of the effect of mean wind
    and stability on a heat island circulation characteristic of an urban
    complex. Monthly Weather Review, 99(12), pp. 919-926.

WMO, 1971:  Guide to meteorological instrument and observing practices.
    World Meteorological Organization No. 8TP3, 4th edition, Geneva,
    Switzerland.

-------
                                                             Section No. 4.1.0
                                                             Revision No.    0
                                                             Datet   17 Sep 89
                                                             Page:    1  of 3

                                  Section 4.1
                   GENERAL ASPECTS OF QUALITY ASSURANCE FOR
                          METEOROLOGICAL MEASUREMENTS
                                    OUTLINE

Section                                                    Pages Rev. Date

4.1.0  OUTLINE AND SUMMARY                                   3    0   9/89

4.1.1  PLANNING FOR A QUALITY ASSURANCE PROGRAM              3    0   9/89

       4.1.1.1  PROJECT DESCRIPTION
       4.1.1.2  PROJECT ORGANIZATION
       4.1.1.3  QA OBJECTIVE
       4.1.1.4  CALIBRATION METHOD AND FREQUENCY
       4.1.1.5  DATA FLOW ANALYSIS
       4.1. 1.6  VALIDATION AND REPORTING METHODS
       4.1.1.7  AUDITS - PERFORMANCE AND SYSTEM TYPES
       4.1.1-8  PREVENTIVE MAINTENANCE
       4.1.1.9  QA PROCEDURES
       4.1.1.10 CORRECTIVE ACTION AND REPORTS

4.1.2  ORGANIZATION OF AUTHORITY AND RESPONSIBILITY          1    0   9/89

4.1.3  QALITY CONTROL GUIDELINES                             2    0   9/89

4.1.4  TRACEABILITY PROTOCOL                                 4    0   9/89

       4.1.4.1  VOLTAGE
       4.1.4.2  WIND SPEED
       4.1.4.3  WIND DIRECTION
       4.1.4.4  TEMPERATURE AND TEMPERATURE GRADIENTS
       4.1.4.5  SOLAR RADIATION
       4.1.4.6  ATMOSPHERIC WATER VAPOR
       4.1.4.7  PRECIPITATION
       4.1.4.8  ATMOSPHERIC PRESSURE

4.1.5  ESTIMATING PRECISION AND ACCURACY                     6    0   9/89

       4.1.5.1  DEFINITIONS
       4.1.5.2  COLLOCATED TRANSFER STANDARDS
       4.1.5.3  OTHER CONSIDERATIONS

4.1.6  SYSTEM AUDITS                                         2    0   9/89

4.1,7  PERFORMANCE AUDITS                                    1    0   9/89

4.1.8  DATA VALIDATION PROTOCOL                              7    0   9/89

       4. 1.8.1  STRIP CHARTS
       4.1.8.2  METHODS
       4.1.8.3  THE EMSL/RTP SYSTEM

-------
                                                            Section No, 4.1.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    2  of  3


4.1.9  QA HEPORTS AND CORRECTIVE ACTION                     2    0   9/89

       4.1.9.1  OPERATIONS LOG AND MAINTENANCE REPORTS
       4.1.9.2  CALIBRATION REPORTS
       4.1.9.3  AUDIT REPORTS
       4.1.9.4  REPORTS TO MANAGEMENT
       4.1.9.5  DISCREPANCY REPORTS

4.1-10 REFERENCES                                           2    0   9/89

-------
                                                             Section No. 4.1.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  3

                     GENERAL ASPECTS  OF QUALITY ASSURANCE
                        FOR METEOROLOGICAL MEASUREMENTS
                                    SUMMARY

    Quality assurance (QA) for meteorological measurements is a relatively new
field.   There are generally two reasons for recording meteorological data.  One
is to learn what the atmosphere is doing, particularly the lower part of the
boundary layer.  The other is to document what the atmosphere is doing.  It is
necessary to find the relevant facts and understand them to learn something .
To document what has been learned may require considerably more data and will
require some assurance the data are correct.  The organizations that need valid
data are the ones which collect it, and they will write their own procedures.
This is how the National Weather Service has handled data collection for
synoptic and climatological applications.  When third party requirements with
the force of law began to need meteorological data for transport and diffusion
modeling and safety analysis, the need for QA was established.

    As with most specialties, HA in general has acquired its own language and
infrastructure.  In this handbook, the goal is to avoid structure which has no
specific value to meteorological data validity.  On the other hand, the goal is
to provide clear definitions, methods and examples which will help produce and
verify valid meteorological data.  Some of the popular sayings or phrases make
valid points.  The book "QUALITY IS FREE" (Crabby, 1979) promotes the idea that
it is really cheaper to do It right the first time.  This concept is easy to
demonstrate with manufactured products where bad products will either cost more
through warranty repairs or lost sales and bad will.  If the "product" is data
or services producing data, an awareness of the ultimate cost of the loss of
data is important.  If no one ever looks at or uses the data, it is a waste
of money to buy, install and operate instruments and recording systems.  Even a
facade of data is not worth the money.  If there is a reason for meteorological
measurement, that reason will provide the basis for estimating the economic
down-side for producing unacceptable data.

    Of course data judged "unacceptable" must have been rejected by someone for
some reason.  The reason for needing the data in the first place will provide
the basis for the economic price which must be 'paid for either not "doing it
right the first time" or for fighting the rejection because it was done right.
Usually the reason for measuring meteorological variables is a government
regulation requirement or a need to combat potential claims of injury.  It can
be argued that the facade of data coming from the instruments on the tower will
satisfy the government regulation requirement.  It has been argued that an
extra nickel spent measuring meteorological variables (or air quality, for that
matter) is a nickel lost to the bottom line profit of a manufacturing plant.

    The assumption upon which this handbook rests is that the pertinent
government regulations and guidance documents will clearly define what valid
data are and how validity is proven.  And further, that those people
responsible for accepting or using the data will require that the data be valid
or rejected.  The expertise necessary for this determination will be found in
this handbook and its references.  This should equally help those who must
collect valid data and those who must accept the claim of validity.  This
assumption clearly moves the purpose for collecting data away from learning and
into the category of documenting.  Documentation needs QA.

-------
                                                             Section No. 4.1.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    I  of  3
4.1.1   PLANNING FOR A QUALITY ASSURANCE PROGRAM
    A formal quality assurance program should be designed into the monitoring
program so that provisions can be made in the measurement system design for
necessary quality control checks and for better monitoring of system
operations.   If these activities are planned and provided for by incorporation
of special readouts, calibration equipment, spares,  and procedures for their
use, then the system is more likely to perform in a satisfactory manner and
deliver valid data.

    The formal plans for quality assurance are presented in a document called a
QA Plan (Lockhart, 1985c and EPA, 1987b).  This plan lists all the
quality-related procedures and the frequency of their use to document the
operation of the instrument system.   The QA Plan contains information under
different headings to organize all the various activities in a logical sequence
and to avoid overlooking an important step.  The specifics of each plan must
relate to the needs of the program,  but the general content elements are the
saae.
       4.1.1.1 Project Description

       This introduction establishes why the documentation of meteorological
data monitoring is needed and why it is important to the organization that
valid data are collected.  It also describes how the data will be used which
establishes the criteria for judging the representativeness of the data.

       4.1.1.2 Project Organization

       The literature of QA abounds with examples of the importance of well
defined organizational structure starting with the organization policy on
quality, endorsed in writing by top management.  This provides the authority to
"do it right the first time."  If the organization has no policy on quality and
if someone at the operations level is given the QA responsibility without
sufficient authority (often the case), the effort may become just another
secondary task which must be done.  This is an invitation to a facade system.
An organization will seldom build a new plant without the expertise of
architects and engineers.  Meteorological data systems are often assembled from
parts picked from catalogs by experts in other fields who do not understand the
routine operating requirements for collecting valid data.  A valid QA Plan is a
structure to encourage and guide organizations toward a successful collection
of needed data.

       4.1.1.3 QA Objective

       This section is the real QA plan.  The first two sections described the
project for which the data will be used and the organization of those who will
participate in the data collection.   This section contains the details of how
the QA program will monitor the collection process with the purpose of
doc-jmenting and defending claims of data validity.

       4.1.1.4 Calibration Method and Frequency

       Calibrations are tests, and adjustments if necessary, to relate the
instrument system to truth or validity.  The evidence of this activity, the
documentation, is the foundation upon which the judgment of data validity must

-------
                                                             Section No.  4.1.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  3

rest.   This section defines in advance how the calibrations will be done, how
often and by whom.

       4.1.1.5 Data Flow Analysis

       This section starts with samples of atmospheric conditions, a rate of
rotation of an anemometer representing wind speed for example, and describes
how the samples are combined into reported values.  The section describes how
these values, perhaps hourly averages, are inspected and judged to be
acceptable or not.  Finally, after validity has been established, the data are
archived in some way to become available for use as the project requires.  An
experienced meteorologist reading this section will know what the data mean and
what data quality control has been applied.

       4.1.1.6 Validation and Reporting Methods

       Section 4.1.1.4 provides points in time at which the instrument
performance is known.  This section describes what criteria are used for any
automatic data inspection programs applied between calibrations and how the
results of such programs are implemented and reported.  If comparisons are made
to other similar measurements, such as a wind speed at a different location or
a different time, this section will document the methodology applied.

       4.1.1.7 Audits - Performance and System Types

       Audits may be required or chosen to add to the documentation some
independent evidence of the performance of the meteorological instruments
and/or the performance of those who are responsible for implementation of the
QA Plan.  This section defines how often performance audits are used to
challenge the measurement instrument system and how often system audits are
used to challenge the implementation of the QA Plan or program.  Also defined
is the type of auditor to be used.  Internal auditors are members of the
organization who are independent from those responsible for collecting and
handling the data.  External auditors are usually outside contractors.  In
either case, the auditor must be experienced in the field of meteorology and
oust be provided support from the operating organization.  Auditing should be
the most positive learning experience for operators and a contributor to data
validity.

       4.1.1.8 Preventive Maintenance

       Some instruments require routine service to assure valid data.  Solar
radiation sensors have glass covers which need to be properly cleaned on some
schedule (daily or weekly), depending on its location.  Tipping bucket
raingages need to be checked periodically for spiders or other insects which
might take up residence in the bucket mechanism.  Anemometers and wind vanes
have bearings which will need service (usually replacement) on some time scale
(quarterly or annually), depending on the environment.  Some dew point sensors
require coating periodically.  All such predictable service points should be
recognized and a preventive maintenance plan described for each of them in this
section.

-------
                                                             Section No. 4.1.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  3
       4.1.1.9 QA Procedures
       Section 4.1.1.2 describes the QA objectives.   This section contains the
details of how these objectives will be met,   A written procedure will both
document how the QA task will be addressed and guide a QA person through the
process.   Procedures are a mechanism for establishing technically correct
aethodology which can be followed by people without the technical background or
experience to write the procedure.   While It Is not practical to use experts to
perform routine tasks, it is necessary to have the expert guidance to follow.
Procedures fill this need.  Procedures should be controlled to the degree that
they cannot be changed without written approval of an expert.  A system audit
checks to see if procedures are being followed as they are written.

       4.1.1.10 Corrective Action and Reports

       Documentation is the main goal of a QA Plan.   General procedures will
retire noting in the Site Log any activity relating to the meteorological
system.   Top management, having established the policy and granted the
authority to "do it right the first time," nesds to be aware of the QA
activities required by the QA Plan,  If a procedure or inspection uncovers a
discrepancy with respect to the clearly written system specifications, a
discrepancy report becomes the message to initiate corrective action.  Top
nanagement needs to see these reports along with the corrective action
statements (usually a part of the discrepancy report form) to know that the
system is In control.  Too often; problems must be visible to top management if
corrective action to the system is going to be Initiated.  Audit reports and
other performance reports are circulated and followed up by procedures
described in this section of the QA Plan.

-------
                                                             Section No. 4.1.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1
4.1.2  ORGANIZATION OF AUTHORITY AND RESPONSIBILITY
    Quality does not mean the best money can buy.  It means that the customer
gets what he needs and expects, no more and no less.   The way to assure a
quality product or service is to first set a top management policy in writing
stating that goal or commitment.  The implementation of such a policy requires
a person who can avoid the conflict of interest of providing the product or
service and Judging its quality before it is delivered.  This usually means a
Quality Assurance Manager.  Once the policy is set and the QA Manager is chosen
the procedures by which quality is assured can be written, usually a QA manual.
The manual might proclaim that every project with a deliverable product or
service will have a QA Plan.

    The authority to establish this kind of organizational structure must be
top management.  During the establishment period, top management must
participate and approve the quality organizational structure and procedures.
Once established, top management can delegate authority to the QA Manager for
operating the department.  A routine feedback from the QA Manager to top
nanagement Is necessary to preserve the control of delegated authority, see
that it is being used effectively and demonstrate to the rest of the
organization a level of importance placed on the quality policy.

    Once the organization is in place, and QA Plans are required, the job of
collecting wind data for a PSD (Prevention of Significant Deterioration)
application starts with a QA Plan.  What are the measurement requirements?  How
will the data be handled?  How will the Instruments be calibrated and serviced?
What procedures will be used?  What outside authority will assure management
that the QA Plan is adequate?  Once all the steps required to gather a year of
valid wind data have been defined, the QA people will monitor the process, help
train inexperienced operators, and build a documentation base to support the
claim of validity for the data once the year is over.  This should include some
outside auditing to add to the documentation an impartial expert opinion of
satisfactory performance, or bring to the attention of the QA people and their
management any problems which might have been overlooked.  The best time for an
independent audit is at the beginning of the data-taking period when the loss
of valuable time can be minimized.

    The principles of this approach to quality are sound and irrefutable.  The
cost is less than any other approach.  All that is required is to figure out,
in. sufficient detail,  what and how the Job is to be done before it begins and
to specify how the job will be monitored to assure satisfactory completion.
Anything less is a gamble which may or may not pay off.  Host organizations do
not like to gamble, but many do.

-------
                                                             Section No.  4.1.3
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  2
4.1.3  QUALITY CONTROL GUIDELINES
    Quality control is a process which operates in parallel with the production
of a product or a service.  There is a gigantic body of literature on the
subject.   Some examples of books are Juran (1979),  Felgenbaun (1961) and Grant
and Leavenworth (1974).  A technical organization,  the American Society For
Quality Control (ASQC) exists for the promotion of quality systems.  ASQC has
committees of volunteers to establish guidelines and standards for the quality
profession.  One such committee is the American National Standards Institute
Accredited Standards Committee (ANSI ASC) Z-l  Committee on Quality Assurance.
A product of this committee is a series of standards (ANSI/ASQC 1987a-e) in
which is stated the need to qualify quality control with an adjective
describing what is to be controlled.  This need is nowhere greater than in
meteorological monitoring.

    In the meteorological literature there are recent papers (Wade, 1987 and
Lockhart,  1988) in which the use of quality control is discussed.   There is a
difficulty with the language used to describe  a task and the perception of
control as related to those working the task.   Assume the task Is the
accumulation of a one year data set of wind and temperature measurements.  The
QA Flan describes the goals and the specifications to which the instruments
must conform.  Purchasing has bought the instruments with a purchase order
requiring conformance to performance specifications and describing how the
conformance will be tested.  A suitable site has been found and a consensus has
been reached as to the representativeness of the site.   Receiving inspectors
accepted the instruments and operators have installed then.  The QA Plan called
for an Independent performance audit at the beginning of the data year which
the system passed.  The QA Plan calls for an inspection of the data on a weekly
basis by the meteorologist or environmental scientist who will be working with
the data.   The QA Plan provides a procedure with which the data Inspector can
coBMinicate in writing with the operators to report questionable data and
receive an answer of special instrument checks.  The QA Plan requires operators
to calibrate the instruments on a six month frequency or when problems are
found.  The QA Plan specifies how the calibrations are to be done and to whom
the reports will be routed.  At the end of the year the data are summarized and
made ready for use in diffusion models.  Where is control and where is quality
control?

    The whole program is controlled by the top management through the QA Plan.
How well the various parts of the organization carry out their responsibilities
is checked on by the QA people or person.  If  the receiving inspectors balk at
performing their service because they are too  busy, top management participates
in the decision to either modify the QA Plan and Policy on Quality or find a
way to accomplish the receiving inspection. Management may choose to gamble
that the instruments are all right and any problems will be uncovered during
the installation and audit.  This is management's prerogative and this gamble
is a pretty safe one.  Without the QA Plan the manager of receiving may make
the decision without the benefit of knowing what the stakes of the gamble might
be.  While the whole project is controlled by  the QA Plan, there is one data
quality control function performed by the meteorologist/environmental
scientist; on a. weekly basis the data are examined and accepted or rejected.
If problems or questions arise, discrepancy reports will be initiated which
operate in accordance with the QA Plan,  The inspector finds that the wind
direction looks too steady and writes it up.  The operator goes to the site

-------
                                                             Section No. 4.1.3
                                                             Revision No.    0
                                                             Date;   17 Sep 89
                                                             Page:    2  of  2


and,  finding that the vane has been removed by a predatory bird, Installs a
spare,  notes the action in the site log and on the discrepancy report,
recalibrates if necessary, and reports back to the inspector closing the loop.
This is a true quality control activity.  The quality of the data is being
controlled by the experienced judgment of the data Inspector who looks back In
the data to find when the vane disappeared, and deletes or flags the data as
missing.

    Another quality control function is the periodic calibration.  This is an
instrument quality control process where adjustments are made as necessary to
keep the instruments "in control."  If more frequent calibrations are used or
if the program goes on for several years, a standard control chart may be used
to visually track the "in control" status of the instruments.  The data quality
control inspector must see the calibration reports and contribute to the
decision about what If anything to do to the data as a result of calibration
findings.  Doing anything to the data requires very careful consideration and
thorough documentation.

    When the organization policy is to achieve the level of quality the
"customer* expects, the whole organization effectively becomes a part of the QA
department.  Various techniques, such a quality circles, may be used to
maintain a high level of quality through broad participation and training.
These techniques also underscore the management's dedication to the quality
policy.  It is only when other criteria, such as departmental profit goals,
enjoy a higher priority than does quality that an independent "watch dog"
organization is required to achieve a published quality policy.  The price for
quality is inversely proportional to its place on the priority scale.  When
quality Is first on the list, the most efficient and least expensive process
can be found for its achievement.

-------
                                                             Section No. 4,1.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  4
4.1.4  IBACEABILITY PROTOCOL
    There is a general practice in QA/QC to use a hierarchy of standards
resting on international standards or those maintained by the National Bureau
of Standards (NBS).  This results in claims of calibrations that are "traceable
to NBS."  While  it is difficult to define traceability in quantitative terms,
there is value in using an authority against which other instruments can be
conpared.  This  section will discuss how this hierarchy relates to
meteorological measurements.

    Most meteorological measurements contain some sensing element which reacts
to the variable  of Interest and the usual transducer outputs of voltage,
current or frequency..  In terns of accuracy, the response of the sensing
element is the most  important and the most difficult to define (see sub-section
2, Specifications, for each variable}.  The measurement of the various
components of the electrical output, including digital code, is straight
forward and subject  to normal methods of calibration and certification.
Protocols for "traceability to NBS" for voltage will be discussed first.

       4.1.4.1   Voltage

       Regular calibration labs maintain transfer standards which are sent to
NBS for calibration.  These in turn are used to calibrate the lab's voltage
sources which in turn are used to calibrate a subject volt meter.  This process
has gone on for  years and is called "traceable to NBS."  Calibration labs check
a volt meter, adjust it  if necessary, and affix a calibration sticker
certifying the meter to be in calibration at the date tested and recommending
re-certification at  a future date {six months to a year).  This traditional
process is entirely  acceptable.

       Manufacturers of volt meters also have transfer standards which they use
to calibrate and certify their products.  Modern digital volt meters or
volt-ohm meters  (DVOM) are very stable in calibration, particularly those of
high quality.  Another method for achieving "traceability to NBS" involves the
comparison of DVOMs.  If one DVOM Is certified as accurate, either by a
calibration lab  or by the manufacturer with a transfer standard, and another
DVOM is placed in parallel across a voltage source, the uncertified one can be
certified by that comparison.  This process is valuable to use in performance
auditing in order to fix the accuracy with which the operator calibrates the
signal conditioners.  The process yields a relative accuracy if neither DVOM is
certified, but it becomes absolute as comparisons are made with certified
DVOMs.

       What is an acceptable error in voltage measurement for meteorological
purposes?  Assume a  measurement system with a full scale output of 1 volt for
wind speed, wind direction and temperature difference.  Assume the DVOMs are on
a range displaying millivolts where the  1 volt full scale looks like 1.000.
How iaportant  is it  if  the  two DVOMs disagree by as much as 0.002 volts?  If
the range for wind speed  is 0.2 to 50 m/s, and the accuracy requirement Is 0.2
m/s, what is that accuracy  requirement expressed in volts?

               50.0  m/s  =  1.000 V
                 0.2  m/s  = 0.2/50 x 1.000 = 0.004 V

-------
                                                             Section No. 4.1.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  4


The disagreement between the DVOMs is equivalent to half the accuracy
requirement.  For all practical purposes a disagreement of this size is not
important, but an auditor would like more information.  Is it a bias or a
random difference?

       A good DVOM, like the Fluke 8060A, specifies its accuracy on the 2.0000
V range as ± (0.04% of reading + 2 digits).  On the 20.000 V range the accuracy
is specified as ± (0.05% of reading + 2 digits).  If the two DVOMs were on the
same output of 0.1000 V (5 a/s for the wind speed example), and if they were on
a range equivalent to the 2 V range stated above, they should each read 0.1000
±(0.00004 + 0.0002) or between 0.09976 V and 0.10024 V.  Truncation of the
•easurenent to fit the display would cause the meters to read between 0.0997
and 0.1002 which would be Interpreted as 4..98 and 5.01 m/s.  If the difference
were as much as 2 ntV (0.1 m/s), it would indicate a bias (calibration) error in
one or both of the DVOMs.  If the DVOMs were each on a 20 V range, they should
read 0.10O ± (0.00005 + 0.002) or between 0,09795 V and 0.10205 V.  Truncation
would force the meters to read between 0.097 and 0.102 which would be
interpreted as 4.85 and 5.10 m/s.  A difference of 2 i»V (0.1 m/s} in the meter
readings could be either a bias error or a random error from the 2-digit
uncertainty..  Switching both DVOMs to the 2 V range would resolve the question.

       Table  4.1.4.1 summarizes the accuracy of the conversion of the
transducer output to voltage output In units of voltage and units of
•eteorology for the 1 volt range example for wind speed, direct Ion and
teaperature difference.

                 Table 4.1.4.1  - Voltage vs.  Met, Unit Accuracy
Variable
Range
Volts
Wind
speed
Direction
4 T

0.000
0.000
0.000
- 1
— -t
- 1
.000
.000
.000
Met. Units
0.0 -
360 -
-5.0 -
50.0
360
15.0
m/s
deg.
C
Accuracy (0.1 % FS)
Volts
0.
0.
0.
001
001
001
Met. Units
0.05
0,36
0.02
m/s
deg.
C
       4.1.4.2  Hind Speed


       Traceabllity to  NBS  has some meaning in the measurement of wind speed.
The National Bureau of  Standards Fluid Mechanics Section operates a pair of
calibration wind tunnels  at their facility in Gaithersburg, MD (Washington,
D.C.).  One can arrange to  send an anemometer to NBS for calibration.  A report
will result which describes the output of the sensor or system (rate of
rotation, or volts) at a series of wind speeds.  NBS states  the accuracy of the
wind speed they use to  be 0.1 mph.  How the user Implements the  test report  is
a different story (see  4.2.12).  If the user is a manufacturer,  the test report
will probably be smoothed by some least square method which predicts speed from
rate of rotation.  The  speed predicted by the rate of rotation of the
anemometer calibrated by  NBS will then be transferred to another anemometer  by
collocating them in another wind tunnel or by calibrating the wind tunnel as an
intermediate standard.  If  new anemometers agree with this  transfer of the
performance of the "standard" anemometer to within some margin of error, the
calibration of the new  anemometer is  said to be "traceable  to NBS."

-------
                                                             Section No. 4.1.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  4


    There is no standard wind.  NBS uses a structure designed to smoothly
control  the air being driven by a propeller-motor assembly.  How smoothly and
uniformly the air flows through the test section is determined by testing,
The wind speed at some point in the test section is calculated from the
•easured pressure difference between the pltot tube and the static pressure,
correcting for air density.   The pressure difference is measured with a
manometer.   Anyone can build a wind tunnel and measure its performance as
accurately as can NBS.  "Traceable to NiS" provides a relative standard of
comparison with absolute errors which are small compared to the needs of the
scientific and Industrial users.

      4.1.4.3 Wind Direction

       "Traceable to NBS" has no meaning as it relates to wind direction
(see 4.2.2.2).

      4.1.4.4 Twnperatureand Temperature Gradients

      There is a hierarchy for temperature much the same as voltage.
Calibration labs and manufacturers maintain sensors with calibrations run by
NBS.  A  user can send an electrical transducer, which has a unique relationship
between  resistance and temperature, to a calibration lab and get a report on
that relationship as determined by the lab's transfer standard.  This
calibration is called "traceable to HBS" because the transfer standard was
calibrated there.  Some concern about how the subject transducer and the
transfer standard are exposed to the "same" temperature is warranted.  The test
•ethod and test facilities are not usually certified by NBS and so the
calibration may not deserve the inferred NBS authority.

       Differential temperature is nothing more than two or more temperature
neasurements taken at different points.  The important calibration for this
variable is one which compares one instrument to the other, a
relatlvecalibration.  Traceability Is not relevant to relative calibrations.

       4.1.4.5  Solar Radiation

       Traceability to an absolute measurement of solar radiation is achieved
by collocated comparisons with secondary standards at organizations such as the
Desert Research Laboratory in Arizona or at a scheduled World Meteorological
Organization (WHO) inter comparison.  An absolute measurement of the intensity
of the direct beam from the sun is made with an Active Cavity Radiometer.  This
instrument is based on fundamental principles.  The cavity "sees" only the
direct radiation from the sun.  The optically black surface in the cavity is
heated by radiation of all wave lengths.  The cavity temperature is accurately
measured and the instrument yields the absolute flux of direct radiation  (D) at
the measurement location.  A global pyranometer with a disc located to shield
the direct beams from the disc of the sun measures the diffuse radiation  (d).
With knowledge of the angle of the sun from the zenith (i), the total global
radiation (G) can be calculated by the following formula.

                    G =s D cos 6 + d

       Secondary standards,  traceable to such an inter comparison, may be used t
calibrate operational pyranometers.  Pyranometers which have been calibrated in
this way may be used as collocated transfer standards in the field.

-------
                                                             Section No.  4.1,4
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  4
       4.1.4,6 Atmospheric Hater Vapor

       It is possible to create an absolute humidity and NBS has a facility for
doing just that.  Atmospheric water vapor instruments, expressing the
conditions as relative humidity or dew point temperature can be calibrated by
NBS in a fundamental procedure requiring only the measurement of length
(volume), mass and temperature.  There are standard methods for creating
relative humidity environments useful for some kinds of instruments.  ASTM
(1985c) describes such a method.  For air quality applications, a collocated
comparison provides adequate accuracy.  If the collocated instrument is a
psychrometer and proper methods are used (ASTM, 1984) for measuring the wet-
and dry-bulb temperatures, traceabllity to NBS might be claimed for the
thermometers used in the psychrometer.

       4.1.4.7 Precipitation

       There are some measurements where traceability to NBS is possible but
not required.  Precipitation measurement is essentially a measure of the volume
of liquid water (including the liquid water equivalent volume of snow) which is
collected by an area bounded by a cylinder.  Calibration may require volumes of
water or equivalent weights.  The accuracy required and expected from
precipitation gages will be well served by the accuracy of commercial measuring
equipment.  Ordinary chemical dispensers such as graduated cylinders and burets
are accurate enough without calibration traceable to NBS.  The measurement of
the area of the cylinder may be made with a commercial ruler or tape, always
keeping in mind the need for quality commercial products for the best accuracy
without extraordinary effort.

       4.1.4.8  Atmospheric Pressure

       Calibration labs create pressures with devices using weights.  It is
possible to use weights with calibration traceable to NBS, but the accuracy
with which the atmospheric pressure is needed for air quality applications does
not require such an effort.

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  6
4.1.5  ESTIMATING P1ECISION AMD ACCURACY

       4.1.5.1 Definitions


       There are about as many definitions of precision and accuracy as there
are bodies devoted to carefully defining these terms.  The definition used here
is found in EPA (1976) on page A17 as follows:

'Accuracy - The degree of agreement of a measurement (or an average of
•emsurements of the sane thing), X, with an accepted reference or true value,
T, usually expressed as the difference between the two values, X - T, or the
difference as a percentage of the reference or true value, 100(X-T)/T, and
sometimes expressed as the ratio, X/T."

       For meteorological purposes, this concept of accuracy is acceptable.
The problem comes from knowing the "accepted reference or true value."  Section
4.1.4 Traceabllity Protocol discusses this problem for all the meteorological
variables of interest.  All data that are
used are averages or means.  The formula for accuracy is

                    n
                   •"•"I
           E » 	 S (X. - T )  - X - T                       (1)
    where
      E  is the average error (accuracy)
      X, is the ith sample of X
      T  is the non-varying true value of X
      n  is the number of samples
      i  is a sample, 1,2,3...n

       Accuracy, the average error, or really the uncertainty  in the value, X.,
has two or three components.  They are bias, conditional bias, and random
error, a statistical expression of a series of which is called precision.
Since, in some cases, bias and conditional bias can be separated, both will be
discussed.

       Precision is defined in EPA (1976) as "A measure of mutual agreement
aaong individual measurements of the same property, usually under prescribed
similar conditions.  Precision is most desirably expressed in  terms of the
standard deviation but can be expressed  in terms of variance,  range, or other
statistic.  Various measures of precision exist depending upon the  'prescribed
similar conditions.'"  This is more difficult to fit to meteorological
neasurements made in the atmosphere because "prescribed similar conditions" are
hard to find.  Parts of the instrument system can be challenged by controlled
environments such as wind tunnels and temperature, humidity or pressure
chambers.  The precision of the measurement can be found, providing it is
larger than the variability of the controlled environment.  Usually it is not

-------
                                                             Section No. 4.1.5
                                                             Revision No,    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  6


larger and what is measured is the variability of the controlled environment.
The definition to be used in this handbook is as follows:
    Precision is the standard deviation of a series or measured values,
    X.,  about the mean measured value, X.  The foi
    of the precision or standard deviation, s, is
X.,  about the mean measured value, X.  The formula for an estimation
                                        .2
                                  / ^f    «yr \
                                  \ ^* * ™" " J                     / ^% \
                                    i                          (2)
                                n - 1
An equivalent formula which can be used for real-time calculations  in modern
data loggers or computers  (Juran, 1979, p 22-7) is
                                                               (33
                                 n (n - 1)
Equation (3) is preferred because it introduces only  two rounding errors rather
than the four introduced in equation (2).  Suppose
X* « X. •+• C where C is some constant value.  Then

        Xt = K'L - C  and  X = X' - C  and  X£ - X « X^ - X' .

Therefore, the standard deviation, s, of a series of  values,  as calculated  by
(2) or (3) is identical to the standard deviation, s, of the  sane series of
numbers, each plus a constant.  If the true value, T» is constant,  the

precision of the accuracy estimate, E, is identical to the standard deviation
of the samples, X,.

       Bias is defined in EPA  (1976) as "A systematic (consistent)  error in the
test results."  It goes on with qualifiers which apply to chemical  labs.  For
meteorology, the definition for this handbook is simply the average difference
between the measured value  X, and  the  true value T.,  (X. - T»),    Using  the

symbols of ASTM  (1984),  the systematic difference  (d) or bias  is  found from the
following equation.

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  6
                         1   n
                         1  ^-
                        •	" •*•  / (X.  —
                         n  L  l
         n

Q ™ mmmnifm  / \f\, ^  i, , /  -™ A "™* A               \^l/
             where

               T,  is the ith sample of the true value of T


It can be seen in these definitions that when the true value, T, does not
change,  the accuracy is the same as the bias.  What is the precision of formula
(D?  For the non-varying true value, the accuracy with which any sample may be
determined is

                     Ei - I ± s                               (5)

and for a variable T

                     EI - d ± s  •                             (6)

When T varies, X may vary in a systematic way.  For example, the case for wind
direction found in 4.2.2.2.2.3 shows a systematic difference, d, of -3.4 deg.
(orientation error) and a conditional bias (potentiometer linearity error) of
about ±0.5 deg. on some wave shape with an amplitude of about  ± 2 deg.  The
accuracy of the vane might be stated as

                    E. = -3.4 ±2.5 deg.  or, removing the

orientation error,  E. «  0.0 ± 2.5 deg.  or, correcting for linearity

at each angle,      E, »  0.0 ± 0.5 deg.  There is little point in

making this final correction in operational programs.  This is  Just an example
of how a conditional bias can be treated to decrease an error or improve the
accuracy.

       4.1.5.2 Collocated Transfer Standards

       If a collocated transfer standard {CTS) is used to estimate accuracy,
any correction which malces the CTS provide values closer to the true value is
worth making.  The statistical method of operational comparability (ASTM, 1984)
is used when a CTS is employed to estimate the accuracy of an instrument
operating in the field.  The analysis requires differences between the
instrument being challenged and the CTS, properly sited.  The differences are
best If the same data logger can sample the challenged sensor,  X , at the same
                                                                el
tine as the CTS, X,, so that the difference,  (X j~X..), can be  stored, squared

aM accumulated.  If this is not possible, the CTS data logger  should form
averages over the same time period as the averages formed in the challenged
system.  Then these averages are used for X  , and X.  ,.  The first calculation

to make is comparability from the following formula:

-------
                                                            Section No. 4.1.5
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    4  of  6
                          /I   "
                      v-z
                       /   n  fel
               (Xai
             where
               X   is the 1th measurement of the subject output
                ell
               X, . is the 1th simultaneous sample from the CTS


The systematic difference,  d, is calculated from (4) substituting
X . and X. .for X.  and T.» respectively.  The estimated standard deviation of

the difference, s. Is calculated from  (8).
                               - d2                           (8)
The minimum sample size,  n ,  required for the calculation of C is given by
equation (9).  Most data loggers sample sequentially.  The time between members
of a data pair to satisfy the requirement of simultaneous measurements must not
exceed one tenth the response time of the Instruments.  The time between pairs
of measurements must be greater than four times the response time of the
instruments to assure sample Independence.
                          , 3s ,2
                                                             (9)
where
  r is one increment of resolution reported by X
                                               tt
For example, a CTS wind vane operating  in a speed range of 2 to 7 m/s
with a delay distance of 2 m would have a response time between 1 and 0.3 s.
If a data logger had an analog to digital conversion cycling
time faster than
0.03 s and if samples were taken no faster than every 4 s, and if the
resolution (measurement and display) of the measuring system were 1 deg, ,  and
assuming a 5 deg. standard deviation of the difference, s, the minimum sample
size would be
          f 3 x 5l2
    n
     o
225 samples,  requiring 15 minutes @ 1 per 4 s.
There cannot be too many samples.  The minimum is specified for a confidence of
99.7% or greater that the estimated mean difference, d, is within one element
of resolution (1 deg. in the example).  At that confidence level, the accuracy
of the estimate increases (error decreases) as the square root of the sample
size.  If the sample size were increased by a factor of 4 to 900, the accuracy
of the estimate would by 0.5 deg.  The values of C and d "found from a series of

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  6


differences are only valid for the range of conditions used for the test.  The
shorter the period of time that is sampled the smaller the. range of conditions
will be.   A reasonable goal for a minimum CIS time period  is 24 hours from the
standpoint of dynamic range variation.

       Assume the CIS will provide the  true value in a test in the free
atmosphere within the limits of the calibration of the CIS to some other
standard,  typically 0.1 m/s with respect to some wind tunnel for speed and 1
deg. with respect to TRUE NORTH for average direction.  The accuracy of the
challenged instrument is the comparability, C.  The bias,  d, provides the
calibration and orientation error.  The standard deviation, s, provides the
irreducible random error or minimum functional precision with which two
instruments measuring the same quantity report when operated using the ASTM
D4430-84,  determines the operational  comparability of meteorological
measurements.  Lockhart (1989) found  the following values  of s for wind speed
and wind direction:

                Wind speed	•	s * 0.2 m/s

                Wind direction	•-	s « 2 deg.

When s is found to be larger than these values, the assumption must be made
that either site bias or a malfunctioning sensor is at fault.  Under those
conditions the calibration error, d,  is also suspect.

       4.1.5.3 -Other Considerations

       The average error, E, calculated over a uniform distribution of X, is
the same as the average difference or bias.  The contribution of the standard
deviation of the difference between X and T  [or the precision error) goes to
zero and the average accuracy is the  bias.  The measurement of meteorological
variables in the atmosphere Is never  really that simple.

       Occasionally there Is confusion  between the word precision and the
resolution of the measurement system.   Resolution is the fineness of the
measurement system, the output of the measurement system or the display of the
output of the measurement system.  A  wire-wound potentiometer in a wind
direction sensor may have a resolution  of 0.355 degrees (1,000 windings of a
wire over 355 degrees).  The circuit  that converts the resistance of the
potentiometer to voltage may have a resolution of 0.1 degree.  The display or
recorded value may show whole degrees.  The whole degree may be truncated from
the output or rounded.  For example,  if the potentiometer  wiper is at the 312th
wire (312 x 0.355 = 110.76 degrees),  the voltage output (312/1000 x 1 volt -
0.3120 volts) has no resolution; the  resolution of its measurement is limited
only by the resolution of the volt meter.  If the system has a digital output
with a resolution of one degree, the  output will be 110 degrees (truncated) or
111 degrees (rounded).  The resolution  of the sensor in this example is 0.36
degrees and the resolution of the system is  1 degree.

       A wind direction system with a resolution of 10 degrees might have a
precision of ± 0.5 deg. (it would take  a lot of samples to prove that
precision).  On the other hand, a wind  direction system with a resolution of
0.1 deg.  might have a precision of ±  3  deg.  (because of hysteresis in the
coupling of the potentiometer to the  vane in the sensor).  Resolution should be

-------
                                                             Section No.  4.1.5
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  6

specified to match the needs of the data application and to provide sufficient
information for data QC.

       Most of the discussion in this handbook includes the sensor and signal
conditioner providing an output.  Good digital data systems degrade the output
so little that they contribute only a small error to the total.  The resolution
and accuracy example 4.1.4.1 applies to any digital system.  There is a two
digit uncertainty in the digit which represents the resolution of the
measurement.  If the analog to digital converter resolves wind direction to 10
deg., the accuracy cannot be better than ± 10 deg.   If the converter resolves
to 1 deg., rounds to the nearest 10 deg. and displays 10 deg., the accuracy of
the average direction may be 3 deg. while the accuracy of a single observation
is t 5 deg.

       When analog recorders" are used, their error must be added to the error
of the measurement.  For some recorders this error can get quite large.  Seldom
considered or specified, for analog recorders which use rolls of chart paper,
is the error caused by expansion and contraction of the paper as a function of
temperature and humidity.  This error added to the resolution uncertainty of
narrow paper rolls marked by a tapping bar (when such recorders are used) can
dominate the error of the system when the analog recorder data are used as
measurement data.

       Random errors identified for each component of a system can be combined
to estimate the total system error by the RSS (root sum square) method.  Biases
or systematic errors cannot be combined in this way.  They must be added
arithmetically.  See Fritschen and Gay  (1979) for further discussion of error
analysis.

-------
                                                             Section No. 4.1.6
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2
4.1.6  SYSTEM AUDITS
    A system audit, as defined in EPA (1976, p A10) is "A systematic on-site
qualitative review of facilities, equipment, training, procedures,
record-keeping, validation, and reporting aspects of a total (quality
assurance)  system, to arrive at a measure of the capability and ability of the
system.   Even though each element of the system audit is qualitative in nature,
the evaluation of each element and the total may be quantified and scored on
some subjective basis."  In short, it is an evaluation of the suitability and
effectiveness of a QA Plan or QA Manual.

    Any audit is most useful when considered as a learning or training
exercise.  Given the newness of the implementation of quality systems to air
quality programs, particularly meteorology, a mechanism for "on the job
training" is useful.  Given the two facts that everyone really wants to do a
good job and almost everyone is a stranger to the concepts of structured
quality systems, an audit is a valuable tool.  There is really little
difference between air quality and meteorology when it comes to system audit
principles.  The short section on system audits in EPA (1976) is slanted toward
air cheaistry projects.  In a very general way, a system audit should include
the following elements for any technical discipline.

       1.   Declared Agenda - The audit should not be a surprise or contain
surprises.   The serious audit is well planned in advance in writing.  The items
to be covered are spelled out.  The agenda is structured with the help of those
to be audited, recognizing the areas where they may need special help.

       2.   Entry and Exit Interview with Top Management - A short introduction
meeting with the authority being audited sets the stage for the cooperation
necessary for success.  Success is defined as improving the audited program
through training and education.  A short exit interview will announce the
findings already discussed with the QA people being audited.  The exit
interview and the audit report should contain no surprises.

       3.   Checklist Structure - The audit should flow along a prepared
checklist of questions, but if time is limited as it usually is, flexibility is
valuable.  Special problems, either found or volunteered should be resolved
even at the expense of failure to finish the checklist.

       4.   Audit Report - The report should be delivered in a timely manner,
certainly no more than 30 days after ths audit, preferably within a week of the
audit.  The report is the important documentation verifying the QA program is
"in control."  It must contain the structure for corrective action with plans
and schedules committed in writing.  An open-loop pledge or a general plan is
likely to get a low priority.  The value of the audit and corrective action
must be clear to the audited organization if the system audit is to be
something other than a paperwork exercise.

       5.   Follow-up - The QA Plan of the organization being audited should
require some form of documentation of the completion of the tasks defined in
the corrective action plan.  Completion of this task closes the loop for the
system audit.

-------
                                                             Section No. 4.1.6
                                                             Revision No.    0
                                                             Date:   17 Sap 89
                                                             Page:    2  of  2

    A short visit to understand the organization, QA Plans and Procedures, and
to meet the QA person is a good first step.  It is difficult to do an effective
system audit without access to all three.

    Since there is a large spectrum of quality programs which might be audited,
it Is important to keep in mind the original reason for the audit and the
objectives and regulations defining the quality program.  One objective must be
to provide the necessary level of quality at the minimum cost as well as to
make the audit process useful and effective.  The goal Is to ensure valid data
with documentation backing up the claim for validity.  This can happen in many
different ways, some of which may not conform to the auditor's concept of a
system.  It is not the role of the auditor to redesign the system, but to
determine if the-system meets the objectives and requirements.

-------
                                                     Section No. 4,1.7
                                                     Revision    No. 0
                                                     Date:   11 Aug 89
                                                     Page:    1  of  1
4.1.7     PERFORMANCE AUDITS
     This handbook will concentrate on performance audits.  The audit methods
for each variable will be described in each of the variable-specific sections.
Tbe purpose of a performance audit is to determine as completely as possible
whether or not the instruments are producing valid data.  It is the
responsibility of the operators to calibrate and operate the instruments so that
they do produce valid data. As mentioned in Section 4.0.0, performance audit
methods may be identical to those used In calibration.  If they are different,
it is expected that the audit method is most comprehensive challenging the
greatest part of the total system.  A complete calibration of an anemometer
requires a wind tunnel.  Most operators do not have access to a wind tunnel and
elect to use the manufacturer's wind tunnel experience as authority for the
aneaometer transfer function.  This practice is generally acceptable where
manufacturers can provide test reports confirming their results.  Statistics on
the distribution of error of samples of production run anemometers with respect
to the generic transfer function are necessary if the manufacturer does not
calibrate each unit during manufacture.  If 100% calibration is a part of
manufacturing process, the method of calibration should be available from the
manufacturer.

     A performance audit may include a challenge by a collocated transfer
standard.  Such data serve to check the transfer function at a few points.  The
uncertainty of the challenge Is usually greater than wind tunnel challenges
where the conditions can be carefully controlled.

     A performance audit on variables such as relative humidity (or dew point
teaperature), solar radiation and atmospheric pressure will usually include a
collocated transfer standard comparison.  This is may be the only way a
challenge can be made to the whole measurement system.

-------
                                                             Section No. 4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89-
                                                             Page:    1  of  7

4,1.8  DATA VALIDATION PROTOCOL

       4.1.8.1  Strip Charts

       Some years ago, meteorological data were recorded primarily on strip
charts.  Average values were estimated by "reading" the charts.  The most
coamon and the most useful strip chart was one where the output of the
measuring system was continuously recorded by means of a galvanometer movement
or a servomotor.  Data QC meteorologists became expert in determining the
validity of the data by examining the strip chart trace, a process necessary
while checking the digitizing of the strip chart data.

       When the digital data logger and computer first appeared, it was the
recommended practice to have strip chart recorders in parallel to the digital
systems.  The strip chart data could be "read" and used to fill the gaps when
the digital system failed, a common occurance in early designs.  Many
aeteorologists found that the strip chart data contained information which was
not present in the digital listing.  The most important Information was the
character of the output during the period of time that the digital system was
sampling and averaging.  One example is the presence and frequency of
potentiometer noise in the wind direction output.  This information Is an early
indicator of potentiometer failure.  Whether or not the digital average was
influenced by this noise could be seen by the comparison of the two outputs,
the strip chart being used as truth.

       Another example shows threshold degradation by the character of the
anemometer trace.  Of course the effect of ice or freezing temperatures on
anemometers and wind vanes could often be seen on the strip chart.  The digital
average value would simply be a number which met the plausibility test but was
erroneous.

       Digital systems have become more reliable, accurate and capable of
on-line processing and at the same time less and less expensive.  The economy
of the digital system pushed the analog recorder to a "back up" role and toward
extinction.  Digital systems show promise of both large memory sufficient to
save one second data samples for time history plots and on-line diagnostic
programs to monitor output patterns for unrealistic variability or lack
thereof.

       Until the technology stabalizes enough to allow a consensus to be
developed and regulatory positions to be taken, requirements for strip chart
data will be a reagon or agency specific requirement.  The one clear fact is
that strip chart data, used as a data quality control tool, will result in a
better data validity protocol.

       4.1.8.2  Methods

       Once data are collected, they should be reviewed to screen out possible
incorrect data points before they are put into accessible storage or passed on
to the user.  While the purpose of a QA program is to avoid the generation of
bad data, it is impossible to do so completely.  Even in the best planned and
best conducted programs, undetected errors can be generated by faulty
equipment,  noisy data transmission lines, faulty key punching, and a myriad of

-------
                                                             Section No.  4.1.8
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  7

other sources.   Filippov (1968) offers a detailed and thorough discussion of
the various possible sources of error.

       In both automatic (ADP) and manual data screening the most obvious
checks should be performed first.  These include such things as being sure that
the data exist and are properly Identified, the forms or files are filled out
properly,  that numbers are in the blocks where they should be, letters are
where they should be, and blanks exist where nothing should be.  This sort of
data editing is a subject unto itself and will not be pursued here.

       Methods of editing or screening meteorological data usually involve
comparison of the measured value with some expected value or range of values.
Techniques for checking the measured value usually fall into one or more of the
following categories:
       1.    Comparison with upper and/or lower limit on the allowed
            range of the data;
       2.    Comparison with known statistical distribution of
            the data;
       3.    Comparison with spatial and/or temporal data fields; and
       4.    Comparison based upon known physical relationships.

       A choice must also be made of what to do with the datura that does not
pass a validation procedure.  Basically there are two choices, eliminate the
questionable data from the file, or flag it for further examination.
Automatically discarding data may be a viable, cost-effective option if the
screening procedure is carefully designed and each datum is not of high value.
Records must be kept of discarded data so the reason for the fault can be found
and corrected.  Flagged data are examined and a decision made on their
acceptability.  If unacceptable, it may be possible to correct them or
substitute a more reasonable value (Reynolds, 1979),  Corrected or substituted
values should be so indicated in the data file, with an explanation of the
substitution available to the user.  Alternatively, data of questionable value
may be kept in the data file under a flagged status, with a notation of why
they are questionable, so that the user can make a decision as to  their
usefulness.  This procedure is of questionable value to most users because the
collecting agency is frequently  in the best position to make a decision on the
data.

       The range test is the most common and simplest test.  Data  are checked
to see if they fall within specified limits.  The limits are set ahead of  time
based usually upon historical data or physically Impossible values.  Some
examples of reasonable range tests are rainfall rate greater than  10 in./hour
or wind direction not between 1  and 360 .  In setting the limits, one must
take into consideration whether  or not the system will select only outrageous,
extreme {i.e., impossible) values usually caused by data handling  errors (such
as wind speeds greater than 100  m/s or less than zero} or just unusually high
(i.e., possible ) values, which  should be examined further.  This  may require a
further decision on just how extreme a value should be flagged.  This decision
should be based on the real impact of using extreme values should  they be  in
error.  Considerations of the cost of incorrect data, the possibility of
correction or substitution, or replacement by obtaining new data should be
made.  Unfortunately, the decision may also frequently be made on  the available
resources of those who examine the flagged data.

-------
                                                             Section No. 4.1,8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  7

                4.1,8.2.1  Comparison with known distributions

                Comparison with known statistical distributions may involve
comparison of means, standard deviations, means of extremes, or higher order
statistics.  For example. Lee and Stokes  (1978) report that their data base
usually had kurtosis of approximately 3 with zero skewness.  Any of their
instruments, then, that showed a marked departure from these values were
considered to be in need of further verification.  (Additional research is
needed to determine whether these or similar criteria could be used in other
areas.)

                Lockhart (1979) suggests  compressing data  into a densely packed
graph where long-term  (week, month, or seasonal) patterns  can be easily seen.
Major departures from  these subjectively  seen patterns can be noted and the
data checked.  Although this method of data verification is usually used to
check a particular data set against a longer term climatology, It can also be
used to check individual values.  For example, one might compare a temperature
reading with the monthly average maximum  or minimum plus or minus
(respectively) two or  three standard deviations.  This technique obviously
depends on a reliable  history or representative measurements being available
from the site and is ineffective for noting significant long-term changes in
the instrument.

                4.1.8.2.2 Comparison with other data fields

                Screening data by comparison with fields of similar or related
data is commonly done  when large amounts  of data are taken and when assumptions
of spatial continuity  of the meteorological variable are physically reasonable.
The most easily visualized example of this is a field of atmospheric pressure
measurements.  Any value can be compared  with those in a large area around it,
either visually, or by numerical interpolation.  Major deviations from the
dominant pattern  (a low pressure reading  in the middle of  a high pressure area)
are not to be expected.  Of course, allowance must be made for meso- and micro-
scale phenomena such as a shortwave or pressure jump area  ahead of a convective
storm.

                Not all meteorological fields can be expected to have the
needed continuity.  Rainfall is a notorious example of discontinuity or
microscale variations.  Wind speed and direction can exhibit continuity on some
spatial scales, but care must be taken to account for the  many effects, such as
topography, that can confuse the issue (See Section 4.0.4.3.2.4).

                Interrelated fields can also be used to screen data.  Rainfall,
for example, is unusual without clouds and high humidity while wind direction
and speed, especially  above the surface layer, are related to pressure
gradients.

                Fields of data in time, rather than space, are also used to
check datum points.  These checks are usually made on rates of change of the
data.  Checks are made both on rates of change that are too high- and not high
enough.  For example,  atmospheric stability is not expected to change by
several classes within an hour.  A wind direction reading, however, that does
not change at all for  several hours may indicate that the  vane is stuck

-------
                                                             Section No. 4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Pmge:    4  of  7
(assuming the wind speed is not zero) or that there is some other problem with
the system.

                4.1.8.2.3 Comparison based on physical relationships

                Screening checks can also be made to assure that physically
improbable situations are not reported in the data.  This kind of check is not
coBMonly used because of the wide variety of conditions that can occur in the
atmosphere under extreme conditions.  These unusual events would frequently be
noted first by some of the statistical or range checks noted above.

               Table 4.1.8.1  Examples of Data Editing Criteria
      Wind Speed:
      Wind Direction:
      Delta Temperature:
      Stability:
      Temperature:
      Dew Point:
      Pressure:
>25 ffl/s (NRC)
>50 .kts (NCC)
>20 kts and doubles at 3-hour observation (NCC)
First 5 hourly values lBC/100m between 10 a.m. and 5 p.m. (TVA)
AT/Az <-l°C/100m between 6 p.m. and 5 a.m. (TVA)
AT/Az >15*C/100m (TVA)
AT/Az <-3.4°C/100B (autoconvectiveHTVAMNRC)
AT/Az changes sign twice in 3 hours (TVA)
A,B,F, or G stability during preclp.  (NRC)
F or G stability during the day (NRC)
A,B, or C stability during the night (NRC)
Change in stability of more than 3 classes
     between 2 consecutive hours (NRC)
Same stability class for >12 hours (NRC)
9 F > mean daily maximum for the month (TVA)
9*F < mean daily minimum for the month (TVA)
> 10*F change in 1 hour at a site (TVA)
First 5 hours within ±0.5*F of next 4 (TVA)
>125°F (NCC)
<-60°F (NCC)
> 108F change 1 hour or 20*F in 3 hours (NCC)
Dew point > temperature (TVA)(NRC)
Dew point change >7*F in 1 hour (TVA)
First 5 hours within ±0.5°F of next 4 (TVA)
>.90°F (NCC)
<-60°F (NCC)
Temp. - dew point >5 F during precip. (NRC)
Temp. - dew point >12 consecutive hours (NRC)
>1060 mb (sea level) (NCC)
<940 mb (sea level) (NCC)
Change of 6 mb or 0.2 inch Hg in 3 hours  (NCC)

-------
                                                             Section No. 4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  7


                Some screening points of this type that are used include
assuring that the dew point is not greater than the temperature, and that the
lapse rate is not greater than the autoconvective lapse rate.  Checks on
stability class versus time (not allowing "strongly unstable" at night or
"stable" during the day) may also be considered in this category.

                Table 4.1.8.1 gives examples of some of the data editing
criteria used by three Federal agencies;  the National Climatic Center  (NCC,
now NCTJC), Kllnt (1979); the Nuclear Regulatory Commission (NRG), Fairobent
(1979);  and the Tennessee Valley Authority (TVA), Reynolds (1978).  Examination
of the table shows some interesting differences that can be ascribed to the
differing missions of the agencies.  Because of their global concerns, the NCC
must allow a far wider range of limits on fields such as temperature and
humidity than does an agency with only local interest, such as TVA.  On the
other hand, the NCC has the data available to do spatial checks over a wider
area than would be possible for many local study situations.  Differences can
also be noted depending on the type of data collection (spot readings once per
hour or three hours versus continuous recordings) and major interests (synoptic
weather patterns versus stability).  Filippov (1968) gives an exhaustive review
of checks used by weather services of many other countries.  The criteria
listed in the table are used to identify data to be edited or challenged for
further review.

       4.1.8.3 The AREAL System

       On the following page is a data validation system recommended for AREAL
to replace the present system for screening meteorological data.  It could be
used to screen data gathered by AREAL, contractors, or state and local
agencies.  The system takes into account the variable nature of AREAL's field
activities.  It does not depend on, or have the advantages of, long-term
multistation network design, nor is it labor intensive.  The basic goal is that
of rapid identification of field problems, with low value assigned to
individual data points, thus allowing the discard of questionable values.
Flexibility is available, however, if an individual project's meteorological
data are judged to warrant a more critical approach.

       The flow of the system is shown in Figure 4.1.8.1.  All data will go
first through a hard copy auditing procedure designed to find data entry and
keypunch errors.  In the hard copy audit, a percentage of data points will be
randomly selected for audit.  A second, independent file of these values, as
well as the hour just before and after the hour, will be created from the
original hard copy.  This file will be compared with the master file and
discrepancies noted.  If there are only a few random discrepancies, these
points will be eliminated from the system.  If there are several, or there
seems to be a systematic pattern of errors, the project office (the office
responsible for gathering and reducing the data) will be notified so that they
can correct and re-enter the data and correct the data entry system.  The data
are next passed through a screening program, which is designed to note and flag
questionable values.  Flagged data will go to the laboratory meteorological
office for review.  There they will either be accepted, discarded, or returned
to the project officer if there is a large amount of questionable data.  That
officer may accept, discard, or correct the data.  The screening values are
given in Table 4.1.8.1,  They offer a combination of range, rate of change, and
physical impossibility checks that are chosen to be reasonably restrictive.  It

-------
                                                            Section No. 4. 1.8
                                                            Revision No,    0
                                                            Date:   17 Sep 89
                                                            Page;    6  of 7


is anticipated that some good data will be flagged, but  that most data handling
and gross instrument failure problems will be caught.
                           Stan
\New Oat
S — N
[Data
\Base
^^•.^^M
V
1

'



V J
a __


1

Zdh/Merge
Program
I
(Umafidated [
«. (


Comparision of Qatg
Selected Data 	
vith Data Obtained
ndependentfy from
Original Source

Screening P

Comparts
Subset of
whhN.W.S
F
™_ — ^
^rogram 	 ........

•
on of
Data Qu>
- Oata na,
I
Final
fa


lit
i

Correct Data
Acquisition/
Processing
Procedures
/\"
ect / flwri^K Protect Office
./ tndicawe o? \1 'Determine if Suspect
X ^curring /^ Data are to be:
X Problem? / Approved.
X / Corrected.
\ / or Deleted.
^y ' 	 1—.,.., nJ
°?^c" Meteorological
Office
\_^ Determine if Data Unresolved
' Approved, Qata
|^ Corrected.
_| er f?9tfftt>et - -.. — — *-
sstionabte
a
Data Corrections,
Approvals, or
Deletions.
1 Validated 1
\ Data File \
     Figure 4.1.8.1
Schematic flow of decisions in the
 AREAi. data validation scheme.
       Data that pass the screening program will go through a comparison
program.  This program will randomly select certain values for manual
comparison with information collected by the National Weather Service.  In the
selection process, one day and one hour will be chosen on which data from all
stations in the network will be audited.  One day in every 20 will be randomly
chosen, and on that day, one hour between 5 and 9 a.m. (EST), will be randomly
chosen.  Data from that hour and day for all stations will be printed out by
the audit program for the manual checks.  The program will also make compressed
time scale plot (20 days of hourly values on one line) for each parameter for
the use of the validators.

       The data generated by the audit programs will first be compared with
National Weather Service data to see if they fit reasonably well with synoptic

-------
                                                             Section No. 4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    7  of  7


conditions prevalent in that area.  The meteorologist will choose the stations
to be used in the verification, and train the data clerks in the subjective
comparison procedure.  All questionable data will be given to the meteorologist
for review as above.  The variables to be checked in this way will include wind
speed and direction, temperature, dew point, pressure, and occurrence of
precipitation.

       Naturally, if the audit checks show a problem with one or more
instruments, an attempt will be made to Identify the time range of that problem
so that all questionable data can be found.  Logs of bad data will also be kept
and used to identify troublesome instruments and other problems.

       This system is suggested principally for AREAL, but may prove a useful
starting place for state and local air pollution agencies wishing to develop a
meteorological data validation procedure.  The suggested system is very
complete and will be evaluated over a period of time.  Changes to the system
may have to be made, depending upon the needs and resources of the users.

-------
                                                             Section No. 4,1.9
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2


4.1.9  QA REPORTS AND CORRECTIVE ACTION

       4.1.9.1  Operations Log and Maintenance Reports

       In all of section 4.1 it has been stressed that the real purpose for a
QA or Quality program is to document data validity and the steps taken to make
that determination.   Any activity which has the potential of affecting the
validity of the data must be reported.  A report usually includes a note in the
station log indicating the time of the operator visit and visual status of the
sensors.   The log is signed by the operator.  If the operator found a problem
which he was authorized to fix, the log would contain the entry to that effect.
If the operator is not authorized to make the repair or does not have the
necessary parts, a maintenance report can initiate the work and the purchase
of the parts to fix the instrument.  When parts are changed, serial numbers or
property numbers should be noted in the log.  If a part does not have a number
(some cup assemblies and propellers do not have numbers), a number of some sort
should be assigned and marked on the part with permanent ink.  The QA Plan
should provide some communication route and method by which the person
responsible for the project and the person responsible for data quality control
(if they are different people) are notified of work done on the system.

       4.1.9.2  Calibration Reports

       Calibration reports are the most crucial documents of a data collection
project.   They are the foundation blocks which uphold the the validity claim.
Quality Control and the routine inspection of the data spans the time between
calibrations.  The calibration reports will show whether or not the system is
"in control."  If the system is always "in control" or operating within the
required tolerance limits stated in the QA Plan and generated by the
application, and the data QC does not have any unsolved mysteries, the data are
valid.  If the calibration shows problems, the report will also show the
corrective action taken or initiated.  The "as-found" and "as-left" readings
are a vital part of the calibration report.  If any data "correction"
(quotation marks used because this Is a very delicate subject) Is applied, the
justification must rest on calibration reports on either side of the data
period and the data in between.  This report requires distribution to the
project leader and the QC Inspector, or at least a sign-off routing.   If
corrective action is Initiated but not completed, a report of completion is
required and has the same routing.

       4.1.9.3  Audit Reports

          Audit reports should confirm the calibration reports.  If they do
not, the assumption is that the audit report is correct.  Whenever a
measurement discrepancy exists, the cause of the difference must be found and
resolved.   If the audit measurements are wrong, the auditor will be smarter
next time and all parties will have more confidence in the calibration reports.
It is the responsibility of the auditor to include a report of the discrepancy
between calibration data and audit data along with the explanation and solution
of the discrepancy in the audit report. It is the responsibility of the
operator to be sure that it is in the report.  The documentation must be
suitable for use in a court of law.

-------
                                                             Section No.  4.1.9
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2
       4.1.9.4  Reports to Management
       Reports to management are of value to maintain the close communication
necessary between the source of authority, top management, and the exerciser
delegated authority,  the QA organization.  Whatever the structure of the
organization, directed effort mist be paid for and planned for through
budgeting.  Top management must know how the quality program is performing its
intended, money saving role.  There are about as many types of meteorological
monitoring programs as there are applications.  One fairly standard one is the
2-level,  60 m tower used at most nuclear power plants.  Crutcher (1984)
provides an insight into costs of a minimum system and an acknowledgement of
the annual costs involved in operations and quality assurance.

"Costs are controlled by the design and reliability of the system, as well as
the marketplace.  Costs given here are approximate 1977 prices for presently
available equipment sufficient to meet the minimum requirements of the Nuclear
Regulatory Commission's Regulatory Guide 1.23 (formerly the USAEC Safety Guide
23).  For the first two years these costs approximate one-third of a million
dollars.   These minimum costs do not include either office or storage space.

    One tower, installation and equipment                    $100,000
    Annual maintenance cost                                    25,000
    Annual cost of surveillance and quality assurance
       (including personnel and supplies,
       magnetic tape, paper etc.)                              50,000
    Annual cost of data listings, etc. based on 15-rain.
       integrating intervals and automatic logging in
       digital form on magnetic tape, 13 parameters
       (channels) to a page, daily summaries                   60,000

The cost of a mobile tower and equipment installation is essentially 50% of the
cost of a permanent-type tower installation.  Other costs remain essentially
the same."

    Of particular interest in this reasonably accurate estimate (ten years ago)
Is the ratio of annual costs to one-time costs, 1.35:1.  Of course, the
requirement of RG 1,23 is for valid data with documentation and quality
assurance.  Smaller systems cost less, but the often neglected provision for
annual operating and QA costs are still necessary if valid data are required,

       4.1.9.5  Discrepancy Reports

       Some systems report discrepancies as a section of another report and
some use a discrepancy report as a stand-alone vehicle to initiate corrections
and report completed corrections.  If it is a stand-alone report, some system
of control is necessary to keep track of open reports and monitor progress
toward completion (called follow-up or needling).

-------
                                                             Section No.  4.1.10
                                                             Revision No,      0
                                                             Date:     17 Sep 89
                                                             Page:      1  of  2
4.1.10.      REFERENCES
ANSI/ASQC,  1987a:   Q90, Quality Management and Quality Assurance Standards -
    Guidelines for Selection and Use. Amer. Soc.  For Quality Control,
    Milwaukee, WI  53203

ANSI/ASQC,  1987b:   Q91. Quality Systems - Model for Quality Assurance in
    Design/Development, Production, Installation, and Servicing.  Amer.  Soc. For
    Quality Control,  Milwaukee, WI 53203

ANSI/ASQC,  1987c:   Q92, Quality Systems - Model for Quality Assurance in
    Production and Installation. Amer. Soc. For Quality Control, Milwaukee, WI
    53203.

AXSI/ASQC,  1987d:   Q93, Quality Systems - Model for Quality Assurance in Final
    Inspection and Test.  Amer. Soc. For Quality Control,  Milwaukee, WI 53203.

ANSI/ASQC,  1987e:   Q94, Quality Management and Quality System Elements
    Guidelines.  Amer.  Soc. For Quality  Control, Milwaukee, WI 53203.

ASm  I984a:  Standard Test Method for MEASURING HUMIDITY WITH A
    PSYCHBOMETER (THE MEASUREMENT OF WET- AND DRY-BULB TEMPERATURES,
    £.  337-84.  Amer, Soc. for Testing and Materials, Philadelphia, PA 19103.

ASTM,  1984b:  Standard Practice for DETERMINING THE OPERATIONAL COMPARABILITY
   OF  METEOROLOGICAL MEASUREMENTS, D4430-84.  Amer. Soc.  for Testing and
    Materials, Philadelphia, PA 19103.

ASTM.  1985c:  Standard Practice for MAINTAINING CONSTANT RELATIVE HUMIDITY BY
    MEANS OF AQUEOUS SOLUTIONS, E 104-85.  Amer.  Soc. for Testing and Materials,
    Philadelphia,  PA 19103.

Crosny, P.  B, , 1979:   Quality is Free.  McGraw-Hill Book Co.,
    ISBN 0-07-014512-1.

Crutcher, H. L., 1984:   Monitoring, Sampling, and Managing Meteorological
    Data, Chapter 4.   Atmospheric Science and Power Production, Darryl
   Randerson, Ed.   DOE/TIC-27601, pp. 136-146.

EPA, 1976:   Quality Assurance Handbook for Air Pollution Measurement Systems,
    Vol. I. Principles, EPA-600/9-76-005.  Office of Research and Development,
    Has. Triangle Park, NC 27711.

EPA, 1987b:  On-Site Meteorological Program Guidance for Regulatory Modeling
    Applications,  EPA-450/4-87-013, Off. of Air Quality Planning and
    Standards, Res. Triangle Park, NC 27711.

Fairsbent,  J, E.,  1979:   (personal communication) Nuclear Regulatory Commission.

Feigenbaura, A. V., 1961:  Total Quality Control Engineering and Management.
    McGraw-Hill, LCCCN 61-11051

-------
                                                             Section No. 4.1.10
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  2

Filippov,  V. V., 1968:  Quality control procedures for meteorological data.
    World Weather Watch Planning Report, No,26, World Meteorological
    Organization.

Fritschen, L. J. and L. W.Gay, 1979:  Environmental  instrumentation.
    Springer-Verlag, NY, ISBN 0-387-90411-5.

Grant, E.  L  . and R. S. Leavenworth,  1974:  Statistical Quality Control, Fifth
    Edition.  McGraw-Hill,  ISBNO-07-024114-7

Juran, J.  M., 1979:  Quality Control  Handbook, Third Edition.   McGraw-Hill,
    ISBN 0-07-033175-8

Klint, W.  E., 1979:  Screening checks used by  the National Climatic Center  for
    meteorological data.  NCC Asheville, NC (unpublished)

Lee, J. T. and J. Stokes, 1978:  Tall tower and aircraft  instrumentation quality
    control procedures - development  and application.  Proceedings 4th
    Symposium on Meteorological Observations and Instrumentation.  Amer.
    Meteor. Sac., pp 3-6.3-6.

Lockhart, T. J., 1978b«.  Data graphics for assessment of  measurement quality.
    Proceedings of the National Conference on  Quality of  Environmental
    Measurements, Denver, Nov. 27-29, pp 123-132.

Lockhart, T. J., 1985c:  Quality assurance of  meteorological measurements.   In:
    "Quality Assurance for  Environmental Measurements," ASTM STP 867.
    J. K. Taylor and T. W.  Stanley  (eds). Amer. Soc. for  Testing and Materials,
    Philadelphia, PA. 253-259.

Lockhart, T. J., 1989a:  Comments on  "A Quality Control Program for Surface
    Mesonetwork Data".  J.  Atmos. and Oceanic  Techno!., 6, 525-526.

Lockhart, T. J., 1989b:  Accuracy of  the collocated  transfer standard method
    for wind instrument auditing.   J. Atmos. Oceanic Technol. ,  6, 715-723.

Reynolds, G. W. and D. E. Pittman,  1978:  The  TVA meteorological data acceptance
    analysis program.  Proceedings  4th Symposium on  Meteorological Observations
    and Instrumentation. Amer. Meteor. Soc., pp 3-6.3-6.

Reynolds, G. W., 1979:  Final acceptance review for  TVA meteorological  data.
    Presented at Quality Assurance  in Air Pollution  Measurement Conference,
    Air Pollution Control Association, New Orleans.

Wade, C. G. , 1987:  A quality control program  for surface mesometeorological
    data.  J. Atmos. and Oceanic Technol., 4,  435-453.

-------
                                                             Section No, 4.2.0
                                                             Revision No.    0
                                                             Date;  Sep  17 89
                                                             Page:    1  of  4


                                  Section 4.2
                QA FOR WIND  SPEED, WIND DIRECTION AND TURBULENCE
                                    OUTLINE
Section                                                    Pages  Rev.   Date
4.2.0  OUTLINE AND SUMMARY                                   4     0     9/89

4.2.1  TYPES OF INSTRUMENTS                                  4     0     9/89

       4.2.1.1  COMMON MECHANICAL SENSORS
            4.2.1.1.1  Mind Speed Sensors
            4.2.1.1.2  Wind Direction Sensors
            4.2.1.1.3  Fixed Component Sensors

       4.2.1.2 SECONDARY EFFECT SENSORS
            4.2.1.2.1  In-situ Sensors
            4.2.1.2.2  Remote Sensing Devices

    4.2.2  SPECIFICATIONS                                   32     0     09/89

       4.2.2.1  WIND SPEED
            4.2.2.1.1  Threshold
                4.2.2.1.1.1  Definition
                4.2.2.1.1.2  Threshold Measurement
                4.2.2.1.1.3  Starting Torque Measurement
            4.2.2.1.2 Accuracy
                4.2.2.1.2.1  Definition
                4.2.2.1.2.2  Measurement of Accuracy
                4.2.2.1.2.3  Application of Accuracy
                4.2.2.1.2.4  Precision
            4.2.2.1.3  Distance Constant
                4.2.2.1.3.1  Definition
                4.2.2.1.3.2  Measurement of Distance Constant
            4.2.2.1.4  Off-Axis Response

       4.2.2.2  WIND DIRECTION
            4.2,2.2.1  Threshold
                4.2.2.2.1.1  Definition
                4.2.2.2.1.2  ThreshoId Measurement
                4.2.2.2.1.3  Torque Measurement
            4.2.2.2.2  Accuracy
                4,2.2.2.2.1  Definition
                4.2.2.2.2.2  Measurement of Sensor Accuracy
                4.2.2.2.2.3  Measurement of Orientation Accuracy
                4.2.2.2.2.4  Expression of Accuracy
                 4.2.2.2.2.5  Precision
            4.2.2.2.3  Delay Distance (Distance Constant)
                4.2.2.2.3.1  Definition
                 4,2.2,2.3.2  Measurement of Delay Distance
            4.2.2.2.4  Overshoot or Damping Ratio
                4.2.2.2.4.1  Definition
                 4.2.2.2.4.2  Measurement of Overshoot
                4.2.2.2.5  Dynamic Vane Bias

-------
                                                             Section No. 4.2.0
                                                             Revision No.    0
                                                             Date:   Sep  17 89
                                                             Page:     2  of  4

Section                                                    Pages  Rev.    Date
       4.2.2.3  TURBULENCE
            4.2.2.3.1  Definition
            4,2.2.3.2  Direct Measurement of Sigma Theta
            4.2.2.3.3  Statistical Summaries

4.2.3  ACCEPTANCE TESTING                                    5     0     9/89

       4.2.3.1  WIND SPEED
            4.2.3.1.1  Threshold
            4.2.3.1.2  Accuracy
            4.2.3.1.3  Distance Constant
       4.2.3.2  WIND DIRECTION
            4.2.3.2.1  Threshold
            4.2.3.2.2  Accuracy
            4.2.3.2.3  Delay Distance and Overshoot
       4.2.3.3  MEASUREMENT SYSTEM

4.2.4  INSTALLATION                                          8     0     9/89

       4.2.4.1  GENERAL CONSIDERATIONS
       4.2.4.2  WIND SPEED
       4.2.4.3  WIND DIRECTION
            4.2.4.3.1  Exposure
            4.2.4.3.2  Orientation
                4.2.4.3.2.1  True Solar Noon Method
                4.2.4.3.2.2  Solar Azimuth Method

4.2.5  CALIBRATION                                           4     0     9/89

       4.2.5.1  WIND SPEED
            4.2.5.1.1  System Accuracy
            4.2.5.1.2  Component Accuracy
       4.2.5.2  WIND DIRECTION
            4.2.5.2.1  System Accuracy
            4.2.5.2.2  Component Accuracy

4.2.6  OPERATIONS, MAINTENANCE AND QUALITY CONTROL           2     0     9/89

       4.2.6.1  OPERATIONS
       4.2.6.2  MAINTENANCE
            4.2.6.2.1  Rout ine and Prevent1ve Maintenance
            4.2.6.2.2  Corrective Maintenance
       4.2.6.3  QUALITY CONTROL

4.2.7  PERFORMANCE AUDIT METHODS                            20     0     9/89

       4.2.7.1  GENERAL CONSIDERATIONS
            4.2.7.1.1  Who
            4.2,7.1.2  What
            4.2.7.1.3  Where
            4.2.7.1.4  When
            4.2.7.1.5  How

-------
                                                             Section No,  4.2.0
                                                             Revision No.    0
                                                             Date:   Sep  17 89
                                                             Page:     3  of  4

Section                                                    Pages  Rev.    Date
       4.2.7.2  WIND SPEED
            4,2.7.2.1  Sensor Control Method
            4.2.7.2.2 CIS Method
            4,2.7.2.3 "W" Anemometers
       4.2.7.3  WIND DIRECTION
            4.2.7.3.1 Sensor Control
            4.2.7.3.2 CIS Method
            4.2.7.3.3 Vertical Wind Direction, <£
       4.2.7.4  TURBULENCE OF er. and 
-------
                                                             Section No. 4.2.1
                                                             Revision No.    I
                                                             Date:  Sep  17 B~
                                                             Page:    4  of  4

               QUALITY ASSURANCE FOR VIND SPEED, WIND DIRECTION
                                AND TURBULENCE
                                    SUMHARY

     This section discusses all aspects of the task of monitoring the wind at s.
particular site with an emphasis on quality assurance.   A background chapter
describes the nature of wind and the kinds of instruments commonly used  to
monitor its speed and direction.  This section describes in detail the
important aspects of the operation of conventional anemometers and wind vanes.
Some discussion of secondary effect sensors is provided but the handbook is nc:
intended to cover these instruments.  The background information and the
detailed information found in the following chapters are necessary for two
kinds of tasks.  One is to execute the responsiblity for the coilection of
valid data.  The other  is to audit or judge how well the first task was
performed within the goals or regulations which caused the measurements  to be
made In the first place.

    Specifications is the longest and in some ways the most detailed section.
The premise is that an understanding in depth of the way the common sensors
work is necessary before purchasing, installing and operating the Instruments-
Specifications set the performance parameters for the instrument or system.
Careful definitions are given along with test methods which will equip the user
to verify or to judge' the work of others who verify conformance to
specification.
    Once the specifications are clearly understood, the process of purchasing
and acceptance testing can be considered.  The contention is that quality
assurance is a vital aspect of defining that which is to be purchased and
verifying the performance of the delivered system.  When the valid system  is i~
hand, the installation  can be planned and implemented.   The important process
of orientation of the wind vane to TRUE NORTH is described in detail.

    Calibration is a foundation on which claims of data validity are built.
This important function may be practiced in a number of phases of the
monitoring program.  This chapter stresses documentation of the calibration
findings and methodology.  The use of the most Inclusive aethods practical in
field conditions is advocated.  Once the calibrated system has been installed,
the routine performance of operational checks, preventive and corrective
maintenance and quality control operations begin.  The documentation of  these
operations provide the  framework, resting on the foundation of calibration, tc
support the claim of data validity.

    Performance audits  add confidence to the documentation that the system is
In control.  Performance audit methods must be the most comprehensive methods
possible for field challenges.  This chapter describes some recommended  audit
methods and audit forms to support  the methods.  The performance audits  and
calibrations provide the data for the estimation of accuracy and precision
described in the following chapter.

    There is not room  to include all the details or background  information
which might be needed or desired.  A list of the references used for this
section is found at the end of the  section.  If the reader needs additional
information or is curious about peripheral subjects, the references will
provide the answers or  a start in search of the answer.

-------
                                                             Section No. 4.2.1
                                                             Revision No.    2
                                                             Date:    17 Sep £9
                                                             Page:     1  of  4
4,2.1  TYPES OF INSTRUMENTS
    There are many ways to detect wind as it passes a point on Earth.   Only
those ways which reference a fixed point (or volume) will be considered in this
handbook.  This class of measurements is expressed in Eulerian coordinates
where properties of the air are assigned to points in space at each given time
(Huschke, 1970).  The other class of measurement is expressed in Lagrangian
coordinates.  It is good to keep In mind that Eulerian measurements are
frequently used in Lagrangian models.  Or,  in other words, monitoring data
measuring wind on a tower are used to estimate where parcels of air move and
how the concentrations of constituents of the parcel change in the process.

    It is necessary to understand just how the measurement is made to
adequately do the following:
       o write procurement specifications,
       « adopt and apply acceptance testing methods,
       <» site the sensors in the representative flow of interest,
       o perform calibration and maintenance services,
       o establish an effective quality control (QC) operation, and to
       « evaluate audits used to estimate precision and accuracy of the data.

    This section will describe how various kinds of instruments work.   The
thoroughness of the description in this handbook will be proportional to the
frequency of use of the instrument in air quality applications.

    Another background point deals with the nature of wind.  There can be no
question about the wind requiring a vector to describe fully a single
measurement.  The vector has direction in spherical coordinates (azimuth with
respect to TRUE NORTH and elevation with respect to a horizontal plane) and
length (speed) along that direction.  It is common for many air quality
applications to deal only with the presumably horizontal flow as measured by
vanes and anemometers.  In this case, the horizontal components of the vector
are expressed as an azimuth angle from which the wind is blowing and the speed
at which the air is passing the point of measurement.  While each sample of
wind requires both speed and direction, it is common to Measure them
separately.  A series of samples may be summarized in different ways depending
upon the application.  The arithmetic mean of the samples is recommended for
many applications (EPA, 1987b).  The standard deviation of the samples is used
to describe the level of turbulence in the air.

       4.2.1.1  Common Mechanical Sensors

                4.2.1.1.1  Wind speed sensors

                Common anemometers are either cup assemblies turning on a
vertical axis or propellers turning on a vane-oriented horizontal axis.  The
cup anemometer is an empirical sensor in that the relationship between the rate
of rotation and the wind speed is determined by testing rather than defined by
theory.  It is a linear relationship, for all practical purposes, above its
threshold non-linearity and through the range of important application.  It Is
an aerodynamic shape which converts the wind pressure force to torque (hence
rotation) because of asymmetrical lift and drag.  Its dynamic performance
characteristics (starting threshold and distance constant) are density

-------
                                                             Section No. 4.2.1
                                                             Revision No.    Z
                                                             Date:    17 Sep 8=
                                                             Page:     2  of  4

dependent but its transfer function (rate of rotation vs. wind speed) is
independent of density.  The cup is not very efficient and creates turbulence
as the air flows through and around it.  The linear speed of the center of a
cup is only a half to a third the linear speed of the air turning it (Mason arc
Moses, 19843.  The cup anemometer is omnidirectional to horizontal flow but
exhibits a complicated reaction to vertical components.  It may Indicate speed
slightly greater than the total speed when the flow is non-horizontal
(MacCready» 1966}.

                The propeller anemometer is a more efficient shape.   The
helicold propeller Is so efficient that its transfer function can be specified
from theory (Gill, 19733.  It creates little turbulence as the air flows mostly
through it, turning like a nut on invisible threads.  The propeller measures
wind speed when it is oriented Into the wind by a vane.  Its errors from
imperfect alignment with some mean vector are small, being nearly proportional
to the cosine of the angle of misalignment.

                In either of these types of anemometers, the rate of rotation
is sensed by some transducer.  Tachometer generators, a.c.  frequency
generators, light choppers and shaft revolution counters have all been used.
It Is important to know how the transducer works if the performance of the
anemometer is to be challenged for a QA purpose,

                4.2.1.1.2  Uind directIon sensors
                The wind vane is perhaps the simplest of instruments.  A fin Is
tied to a vertical shaft such that when force is applied to the area by the
wind, it will turn the shaft seeking a minimum force position.  The
relationship of the shape, size and distance from the axis of rotation of the
fin to the bearing assembly and transducer torque requirements determines the
starting threshold.  These attributes of the fin area along with its
counterweight determine the dynamic performance characteristics of overshoot
(damping ratio) and delay distance (distance constant) of the direction vane.
While its equilibrium position is insensitive to density, the dynamic response
characteristics and threshold are density dependent.

                Vane design is of little importance if the average wind
direction is all that is required.  If turbulence parameters are of interest,
as they usually are or should be, the design of the vane becomes important.
The vane transducer Is usually a potentiometer, but synchros, shaft encoders,
capacitors and Hall effect devices have been used.  It Is fairly common to find
the range of the sensor to be "540 degrees" rather than the physically true
360 .  The reason is related to the problem of a continuous range (a circle)
with a discontinuous output (0 to n volts).  It is Important to know how the
transducer works if the performance of the wind vane is to be challenged for a
QA purpose.

                A special direction vane is the blvane which has the vertical
range of ± 45° to 60°in addition to the full azimuth circle.  The additional
range brings with it the need to neutralize gravity by having a perfectly
balanced vane assembly.  Blvanes can be conditionally out of balance, such as
happens when dew forms and then evaporates from the tail fins.  The effect of
this Imbalance on threshold and performance is complicated.  Horizontal vanes

-------
                                                             Section No. 4.2.1
                                                             Revision No,    1
                                                             Date:   17 Sep 89-
                                                             Page:    3  of  i

can be designed to be stable in the horizontal even when slightly out of
balance.  The effect of this design is to add the vane horizontal restoring
force to the wind force, again a complication.


                4.2.1,1.3 Fixed component sensors

                Propeller anemometers exhibit something like a cosine response
to a wind along some line other than the axis of the propeller.  The degree
with which this response represents a cosine is a function of the design of the
propeller.  If the cosine response is perfect, the fixed propeller accurately
reports the component of the wind parallel to the axis of the propeller.  If
three propellers were located on fixed X, Y and Z axes, the three outputs would
define the components of a three dimensional wind vector.  From a QA
perspective, the accuracy of the wind speed and direction data are related to
the determination of the component errors and the algorithms used to correct
for them.  Often ill-defined are the errors from the interference of one
propeller on another and the errors when the beyond-threshold-nonlinearity
speed has not been reached.  It is the speed of the component parallel to the
anemometer axis that the propeller responds to, not the total speed.  A 5 m/s
wind with a 5° up angle and 5° off the Y axis will provide "a 0.44 m/s wind for
the W and X propellers.  A 50% error in the X propeller because of threshold
nonlinearity would cause an insignificant 0.014 m/s error in the wind speed aad
a 2.5  error in wind direction.  A 50% error in the W propeller for the same
reason would cause a 50% error in the W component (0.22 m/s reported rather
than the true 0.44 m/s).

       4.2.1.2  Secondary Effect Sensors

                4.2.1.2.1  In-sltu sensors

                Several meteorological instrument books contain information on
a variety of wind Instruments.  See Mason and Moses (1985) and Middleton (1953)
for greater depth and variety.

                The three component sonic anemometer is considered in some
circles as the standard for wind measurement.  For those applications where the
contribution of small eddys is Important, it is an excellent choice.  As with
many of the secondary effect sensors, it Is a research tool requiring
considerable attention from the operator.  It is not a good choice for routine
monitoring.  It has its own set of error sources when it Is used for measuring
long-term (tens of minutes) averages and standard deviations.  Single component
sonic anemometers deployed as W sensors may become the standard for this
difficult measurement.  The secondary effect for this instrument Is the
transport of sound waves in the air.

                The hot wire or hot film anemometer is also a research tool
which measures the wind speed component perpendicular to a heated cylinder.
The secondary effect for this instrument is the removal of heat by the air
measured by the current it takes to replace the heat.  There are some new
designs of this type of instrument which are intended to be monitoring
instruments.

-------
                                                             Section No, 4,2,1
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     4  of  4
                4.2.1.2.2 Remote sensing devices
                There are two Doppler shift instruments which measure wind
remotely by analyzing the return from transmitted energy pulses.  The most
important for the boundary layer applications is the acoustic Doppler (SODAR)
sounder.  It depends on the back scatter from small temperature differences
that tag the air for motion measurement.  The other Doppler uses
electromagnetic energy to measure winds through the troposphere.  There are
also some Doppler applications using lasers as the energy source.,  These
systems are complicated,  A QA effort related to systems of this type will
require special study of the system and ingenuity to find other ways of
measuring what they are measuring.  In the case of the SODAR which is being
used in monitoring applications for air quality programs, a reasonableness test
with another instrument capable of measuring winds aloft is more likely to
produce useful information than is a challenge designed to report the accuracy
of the SODAR.

-------
                                                             Section No. 4,2,2
                                                             Revision No.    j
                                                             Date:   17 Sep 85
                                                             Page:    1  of 32
4.2.2  SPECIFICATIONS
    The purpose of defining specifications is to give unambiguous meaning to
the terras used by all those who are concerned that the instruments and systems
selected and operated will meet the needs of the application or project.  This
starts with procurement specifications and ends with supporting claims of data
quality. These specifications provide the basis for receiving inspection and
testing.  The wind is the most important variable to be measured and its
specifications are the most complicated.  Specifications discussed here will
also include some aspects of the measurement system.

    Project and application requirements vary.  To make this handbook as
specific as possible, the examples used will be consistent with those presented
in the On-Site Meteorological Program Guidance for Regulatory Modeling
Applications (EPA, 1987b).  The specifications will be discussed in order of
their importance and then summarized at the end of the sub-section.

       4,2.2.1  Vind Speed

                4.2.2.1.1  Threshold

                     4.2.2.1.1.1 Threshold definition

                     One of the keys to a good wind sensor is a low threshold.
The threshold is also the one performance characteristic which will certainly
change with time because of bearing degradation.  There is no standard
definition of threshold so different manufactur-ers may apply different tests
to establish their threshold specifica-tion.  Absence of a standard or
definition of the specification makes it difficult to specify a meaningful
value.  The following definition comes from Standard Test Method for
DETERMINING THE PERFORMANCE OF A CUP ANEMOMETER OR PROPELLER ANEMOMETER (Draft
6) (ASTM, 1985):
         "Starting threshold (U ,m/s)—the lowest wind  speed at  which a.
          rotating anemometer starts and continues to turn and produce
          a measurable signal when mounted in its normal position."
A starting threshold specification, 0.5 m/s for example, should include a
footnote describing the meaning of the specification. In the example above, it
might say:  0.5 m/s (1)
          (1) "as determined by wind tunnel tests conducted on production
          samples in accordance with ASTM D22.ll test methods."

                     Ail rotating anemometers are non-linear as they go from not
turning to turning at a rate predicted by their linear transfer function.
Note that the definition does not require linear output at threshold, only
continuing turning and measurable signal.  If the manufacturer provides an
accuracy  specification which is independent of speed, the presumption is that
the accuracy specification  is met at threshold.  Consider a hypothetical cup
with a transfer function, i.e., the relationship between rate of rotation and
wind speed, as follows:
                                U - 0.2 + 1.5 R
                        where U is wind speed (m/s) and
                          R is rate of rotation  (rps)

The transfer function would have been found by using a least squares fit
(linear regression) to wind tunnel data.  The ASTM method uses the wind speeds
well above the starting threshold to avoid bias from the non-linear threshold.
In Figure 4.2.2.1 the lowest 2 m/s of the hypothetical performance curve is

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page;     2  of 32
shown along with the contribution of the offset to the system output.  The
variable part of the transfer function (U ~ 1,5 R) coming from the cup rotation
is shown theoretically as the straight line from 0,2 m/s to an output of 1.8
m/s when the wind speed is 2 n/s.  The triangles show the actual output from
the cup rotation.  They start to turn at 0.3 m/s (threshold] and reach the
theoretical line at about 0.8 m/s.  The parallel line through the origin simply
adds the constant offset to the cup rotation output.  The measurement error is
the difference between the diamonds in the figure and the ideal straight line.
It starts at +0.2 m/s, goes to -0.1 m/s at 0.3 n/s,  and then gets smaller as
the nonlinearlty of the threshold decreases.

                     The offset is defined either by the linear regression or
by the arbitrary choice of the manufacturer.  If it is the former, the starting
threshold will always be larger than the offset.  If it is the latter, the
starting threshold may be either side of the offset.  The manufacturers of the
common small three cup anemometer often set an offset voltage in their signal
conditioner as shown in Figure 4.2.2.1.  For this hypothetical cup, the offset
voltage is critical to its meeting the accuracy specification discussed in
4.2.2.1.2.  Sensitive propeller anemometers have a much smaller offset because
they develop more force (torque) per m/s. Some offsets are so small that there
is no advantage or need to use an offset voltage.  See Baynton (1976) and
Lockhart (1977} for further discussion of the errors of rotational anemometers,
particularly at the threshold.

                Cup  Anemometer  Performance
                                 Threshold Analysis
             Hypothetical Transfer Function
             U (m/s) « 0.2 + 1.5 R (rps)
              From Cup Rotation and Offset Voltage

                      Ideal

               O     Actual
                                             From Cup Rotation

                                                      Theoretical

                                                V     Actual
                0.2    0.4    0.6    0.8    1.0    1.2    1.4    1.6    1.8
                                  Wind  Speed (m/s)

Figure 4.2.2.1  A hypothetical cup anemometer threshold analysis.

-------
                                                             Section  No.  4,2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     3  of 32
                     4.2,2.1.1.2   Threshold Measurement

                     There  is  only one way to measure starting threshold.   It
requires a wind tunnel capable of  accurate operation below  1 m/s. One  standard
methodology is defined in ASTM (1985) and described in Lockhart  (1987).
However, it is possible  to  estimate the  starting threshold  by matching the
torque which is required to keep a cup or propeller from turning at a  known
wind speed (in a wind tunnel)  with the starting torque of the anemometer
bearing assembly.  Lockhart (1978)  provided the torque relationship as a
function of wind speed for  four anemometer shapes.

                     Table  4.2.2.1  contains values calculated with these
data by using the relationship
                            T = K u2
          where T  is torque (g cm /sec  )
                u  is the square of the  wind speed (m/s)
                K  is a  constant for the aerodynamic shape  (g)
The values in the table  were calculated  from this formula using
the K values from Lockhart  (1978).

                Table 4.2.2.1  - Torque Developed vs.  Wind Speed
Speed
(m/s)
0.1
0.2
0.3
0.4
0.5
1.0
cup
#1
(g-cra)
0.014
0,056
0.126
0.224
0.350
1.4
cup
#2
(g-cm)
0.027
0.108
0.243
0.432
0.675
2.7
prop
#3
(g-cm)
0.049
0.196
0.441
0.784
1 .225
4.9
cup
i4
{g-cm)
0.148
0,592
1.332
2.368
3.700
14.8
#1 Teledyne Geotech 170-42 (20.3 g) K* 1.4
#2 MRI Model 1022 (48.3 g) K= 2.7
#3 R.M. Young Model 21180 (9.7 g) K= 4.9
#4 MRI Model 1074 (186.8 g) K=14.8
The torques listed are those acting on the sensor when the sensor is restrained
in a wind field at the speed listed.  If the sensor bearing assembly has a
starting torque less than the torque provided at that speed, and the restraint
is removed, it will start turning.  The torque watch used for the low speed end
of the wind tunnel tests was a Waters Model 366-3 with a range from 0.003 oz-in
(0.216 g-cm) to 0.03 oz-in (2.16 g-cm).  To convert oz-in to g-cm, multiply by
72.
                     The method of using a starting torque measurement to find
the sensor starting threshold will become standard only with the publishing of
K constants by the manufacturers.  One manufacturer (R.M. Young Co.) provides
the K value for anemometers.  These values are shown in Table 4.2,2.2.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    C
                                                             Date;   17 Sep 89
                                                             Page:    4  of 32
                      Table 4.2.2.2 - Anemometer K Values
Type
Polypropylene Cupwheel
Polypropylene Propeller
Polystyrene Propeller
Polystyrene Propeller
Model
No. 12170C-100CB1
No. 08234- 18x30cm
No. 21282-19x30cm
No. 21281-23x50cm
K
1.4
2.5
3.6
5.0
                     4.2.2.1.1.3 Starting Torque Measurement

                     The starting torque of an anemometer bearing assembly will
increase in time because of wear and dirt.  The starting torque, with the cup
assembly or propeller removed, can be measured. Starting torque measurement is
simple in concept but sometimes difficult in application.  An experienced
meteorological instrument technician can tell if a bearing assembly is in need
of service by simply feeling the shaft or rotating or spinning the shaft and
listening to its sound.  The trouble with this practice is that it is not
quantitative.   It works for field servicing instruments but does not provide
documentation suitable for a quality control program.  Another qualitative
practice is to roll the sensor slowly over a smooth horizontal surface watching
                                            the shaft not turn as the sensor
                                            turns around it (see Figure
                                            4.2.2.2).   Set screws and other
                                            asymmetries apply a torque which
                                            keeps the shaft from turning while
                                            the sensor moves around it.  If the
                                            applied torque could be measured,
                                            this method would be quantitative.

                                                  The measurement of the
                                            starting torque of the bearing
                                            assembly provides only an
Figure 4.2.2.2 Climatronics F460 torque     approximation of the starting
               test for speed sensor        threshold of the anemometer,
particularly cup anemometers.  The absence of the cup weight may lower
the starting threshold of the cup bearing assembly but there Is no evidence
that this is an important consideration.  At this point in time there is no
better way to estimate and document in the field and in units of wind speed
this important specification, the starting threshold of the anemometer.

  The direct measurement of starting torque requires some device which can
apply a known torque.  The most common, perhaps, is the Waters Torque Watch.  A
model 366-3 is shown in Figure 4.2.2.3 applied to a Climatronics cup anemometer
sensor.  The measurement requires some degree of care and skill.  The torque
watch has a square shaft which fits into a square hole in the connecting
fixture.  The torque watch is turned while holding its shaft in line with the
anemometer shaft, without end loads.  The indicator is watched and when the
shaft turns the maximum reading is recorded.  This process needs to include at
least one full turn of the anemometer to be sure the maximum friction in the
bearing assembly is encountered.  The torque watch measures either clockwise or
counterclockwise.  Use only the rotation sense required by the cup assembly or

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of 32
propeller. The range of the torque watch may not be as sensitive as one would
like.  If the model 366-3 is used on anemometer ttl in Table 4.2.2.1,  the
threshold (0,003 oz-in or 0.216 g-cm) will not measure equivalent speeds below
0.4 m/s.   If the torque watch turns the shaft
without reaching the lowest scale point, all
that can be said is that the starting threshold
of the anemometer is less than 0.4 m/s.

     Another torque watch is the GM-CM Torqmeter
781 with a 0.1 to 2.1 g-cm range (0.001-0.029
oz-in), shown In Figure 4.2.2.4 mounted to a
Teledyne Geotech  1565C  wind direction sensor.
A third torque measuring device is the simple
Torque Disc, model 18310 made by R.M. Young Co. ,
shown in Figure 4.2.2.5, This is a fundamental
device which does not need expensive
calibration.  Weights (screws) are attached at
distances from the center of rotation.  The
force of gravity provides g-cm torques at the
center of rotation of the intentionally out of
balance disc.  The shaft being tested must be
horizontal and symmetrical in mass.  A cup
anemometer shaft which does not turn while the
sensor is slowly rolled along a flat surface
will not work with the Torque Disc.  The g-cra
torque applied equals the weights and distances      Figure 4.2.2.3 Vaters
when the weights are in the same horizontal                        Torque Watch
plane as the shaft.  Calibration results from
weighing the weights and measuring the distances.   An appropriate Interface
fixture would allow the Torque Disc to be used to calibrate a torque  watch.
Figure 4.2.2.4 Gm-Cm
Figure 4.2,2.5 Young Torque Disc

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep £9
                                                             Page:    6  of 32
 There are several ways to measure torque but the available
instrumentation to make the measurement is limited.   Figure 4.2.2.6 is a
collection of spring-type torque watches, spring scales and circular discs
capable making torque measurements within narrow ranges and specific
orientations.  It is necessary to become familiar with these devices and how
they are correctly used.

                                                         A - Waters Torque Wetch
                                                             Model 366-3
                                                             (0.2-2.0 g-cm)
                                                         B - Waters Torque Watch
                                                             Model 651X-3
                                                             (18-360 g-cm)
                                                         C - GM-CM Torqmeter
                                                             Model 781
                                                             (0.1-2.1 g-cm)
                                                         D - Young Torque Disc
                                                             Model 18310
                                                             (0.1-15 g-cm)
                                                         E - Haldex AB
                                                             Gram Gauge
                                                             (1-10 g)
                                                         F - Young Gram Gauge
                                                             Model 18330
                                                             (0-10 g)
Figure 4.2.2,6  Various Torque Measuring Devices

                4.2.2,1.2 Accuracy

                     4.2.2.1.2.1 Definition

                     The classic definition of accuracy is the comparison of a
measured value to a true value expressed as a bias term plus or minus a randoa
uncertainty (precision).   The bias term may be conditional with respect to the
best fit straight line; it may vary with wind speed or angle of attack.

                     When accuracy is specified, the kind of true value to be
used to test the accuracy claim must also be specified.  Usually the buyer
expects the "true value"  to be the wind speed where the anemometer is sited.
The manufacturer expects the "true value" to be the near laminar flow of a
calibration wind tunnel.   Some auditors expect the "true value" to be the
output predicted by the transfer function when the anemometer is rotated at a
known rate.  Let us label the kind of accuracy as follows:
       A(l) - accuracy with respect to the horizontal component
              of wind speed at the sited location,
                A(la) - instrument response
                A(lb) - siting representativeness
       A{2) - accuracy determined in a wind tunnel, and
       A{3) - accuracy of conversion of rate of rotation to output.

                     A(3) is the easiest to measure and represents most of the
claims for data accuracy from audit reports.  It requires, usually, a measure

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    7  of 32


of the offset voltage  (or equivalent m/s} and the combination of offset and
output from the rate of  rotation  to output converter.

                A(2) requires a calibrated wind tunnel and it is a more
difficult and more expensive accuracy to determine.  It does provide a check on
the manufacturer's generic  transfer function plus the variation in production
individuals represented  by  the individual being tested.  The wind tunnel
calibration test provides a specific transfer function which can be used to
change the signal conditioning to get the smallest error of an A(2) type for
that specific system.  Serial numbers should be recorded with the test results.

                     Some manufacturers do not identify their cup assemblies or
propellers with numbers.  Assume  an A{2) accuracy is found in a wind tunnel
test and compared to the generic  transfer function used for A{3) operations.
Assume the test shows  that  the A(3) value is uniformly 10% low.  Assume the
operators took a year  of data with AC3) accuracy tests showing insignificant
errors.  The error to  the A(2) level is a 10% bias and the data can be
corrected for the year.  This action requires either evidence or good reason to
believe that there was no physical change to the cup assembly over the year and
that the individual tested  was the one used during the year.

                     There  are still unknowns of the A(l) type to consider.
These are usually conditional biases and often impossible to define.  They may
be recognized and their  impact estimated.  There are two types of these errors.
One is the consequence of the anemometer design in the flow field it is to
measure., A(la).  The  other is a  result of assumptions of representativeness,
                      The discussion of A(la) errors requires an understanding
of conventions for  the use of  u,  v and w.   In  traditional diffusion
applications, the statistics for  wind representing a period of time refer to u
as the speed of  the horizontal component along the direction of the mean wind,
v as the speed of the horizontal  wind component perpendicular to the mean
direction, and w as the speed  of  the vertical  wind component.  Another
convention applies  to fixed component anemometers such as the UVW propeller
array.  Here, the U is the east-west component of the wind in a Cartesian
coordinate system (a west wind is positive); V is the north-south component of
the wind (a south wind is positive); and W  is  the vertical component of the
wind {upward moving wind is positive) (Stull, 1988).

                      MacCready (1966) characterized errors in anemometers when
operating in a turbulent flow.  A cup anemometer has a u-error because of a
different response  "constant"  to  an increasing speed than to a decreasing
speed, so-called overspeeding.  With modern sensors, this is usually a small
error of a percent  or two, depending on sensor design and height above ground.
A cup anemometer has no v-error since it is insensitive to changes in
direction, but it does have a  w-error caused by non-horizontal flow.  This
error can easily be 10% and larger with some designs (Lockhart, 1987).  A vane
oriented propeller  will have small v-errors and w-errors from misalignment.
These will be small because the propellers  respond nearly as the cosine of the
misalignment angle,  2% for a 10 degree misalignment. The u-error is too small
to measure for light weight helicoid propellers.  These are all A(la) errors
and they vary as the wind varies.

                      A(lb) errors deal with the assumption is that the
anemometer is measuring what the  true wind  would be at the point of measurement

-------
                                                             Section No. 4.2,2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    8  of 32

if the anemometer were not there.  This is a question of representativeness and
not Instrumentation but it can have a large impact on the question of data
accuracy.  The influence of any supporting structure can bias the flow which
the anemometer faithfully measures.  If the assumption goes further to equate
the measurement to its physical height above ground, and if the anemometer is
mounted on a 2 m pole on top of a  large 8 ra building, the bias with respect to
a 10 m flow over a flat field will be the fault of the building.  These errors
are of the A(lb) type.  If the pole is on the edge of the building, the
distortion of the building will provide non-atmospheric errors of the A(la)
type to be combined with the A(lb) type.  These types of errors are very
difficult to define and virtually  impossible to correct.  Data from an
anemometer mounted in a questionable site, after the A(2) and A{3) errors have
been calibrated out, could be compared with data from a vane oriented propeller
anemometer mounted in a space where the subject anemometer is assumed to
represent.  The difference in these collocated measurements may be used to
estimate the magnitude of A(l) errors.  The A(2), A(3) and A(la) errors are the
ones to concentrate on minimizing.  See 4.2.4.1 for siting guidance.

                     4.2.2.1.2.2  Measurement of accuracy

                     The accuracy of an anemometer is found by comparing its
output to the known speed in a wind tunnel.  A calibrated wind tunnel has
uncertainties associated with its operation.  These incliide instrumentation
errors in measuring the wind speed in the tunnel when it is empty (0.1 raph or
0.05 m/s in the NBS wind tunnel above 2 mph) and the inhomogeneity of wind
speed in the test section away from the boundary layer  (a function of the
tunnel design).  The turbulence level in the wind tunnel test section should be
homogeneous across the test section with most of the energy in eddy sizes which
are small compared to the size of  the anemomater.  When this is true, and it
usually is, turbulence does not influence the calibration.  Fluctuations in the
tunnel speed can be thought of as  long wave length longitudinal turbulence.
This "turbulence" can influence the calibration without careful measurement
synchronization and time averaging.

                     When an anemometer is placed in the test section for
calibration, consideration must be given to blockage errors, which are
dependent on the ratio of the size of the Instrument to the size of the test
section.  Also Interference errors, which are dependent on the placement of the
anemometer with respect to the wind tunnel instrumentation need to be
considered.  Small calibration wind tunnels may themselves be calibrated with
an anemometer which has been run in the NBS wind tunnel.  It is also a common
practice to run two anemometers side by side, one of which has an NBS
calibration curve.  It is prudent  to reverse the positions from time to time to
verify test section homogeneity. It is not reasonable to expect such
calibrations, even though they are "traceable to NBS" by some definition, to
have an accuracy better than 0.2 mph {0.1 m/s).

                     While a calibrated wind tunnel is  the recognized standard
method for calibrating an anemometer, a fundamental (but not very practical)
calibration is possible by moving  an anemometer over a measured length in a
measured period of time through still air (Lockhart, 19S5b and Stearns, 1985).
Most manufacturers have samples of their products calibrated by NBS to
establish for their design a generic relationship between wind speed and rate
of rotation, measured by counting  pulses, frequencies or output voltages.  This

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    9  of 32


relationship is then used as the transfer function to define what the signal
conditioning electronics or other output devices require to express the
measured rate of rotation in units of wind speed.  Some manufacturers test and
adjust each cup wheel or propeller to fit the generic relationship within some
error band.
                     Baynton (1976) discusses the calibration of anemometers
and shows the results of tests of 12 different kinds of anemometers in the
National Center for Atmospheric Research (NCAR) wind tunnel.  He compares his
calibration to the manufacturer's calibration or generic relationship.   Except
for the Aerovane,  which probably was too large for the NCAR tunnel, the
difference was within ±3%,  He also discusses the difference between an
anemometer transfer function which goes through the origin, of the form
                              Y = bX,
and the transfer function with an offset or threshold, of the form
                              Y • a + bX.
Table 4.2,2.3 summarizes data from his Table 1 and Table 5.
            Table  4.2.2.3  Wind Tunnel Test Results From Baynton

Type of Anemometer
Gill 4-blade helicoid propeller
Gill 3-blade helicoid propeller
Aerovane helicoid propeller
Taylor Biram* s propeller
Casella Sensitive
Thornthwaite
INSTAAR t
Climet 011-1
TechEcology %
Gill 3-cup
Electric Speed
Bendix Totallizer Model 349
MR! Model 1074 ft
* a is not significantly greater
t Institute for Arctic and Alpine
a
(.m/S)
0.073*
0.011«
0.233
0.145
0.467
0.331
0.316
0.265
0.275
0.250
0.610
0.588
0.087»
than zero
b
(m/rev. )
0.309
0.487
1.356
0.255
1.404
1.476
1.597
1.382
1.391
1,057
2.728
2.605
2.314

Research, Boulder,
E

1.03
1.03
0.93
1.00
0.98
0.97

1.01


1.03
0.97


Colo.
$ Analysis of NBS data provided by L. Petralll
tt Data from Lockhart (1977)
E is the ratio of the test result

to the mfg


. ' s calibration.
Baynton lists the value of "a" for the 12 anemometers he tested, which ranged
from nearly zero (0.01 m/s) to 0.6 m/s.  One purpose of his paper was to
caution users of "wind run" instruments of the errors associated with ignoring
"a."  Wind run is generally used to describe those anemometers which count
shaft rev-olutions over a long period of time or are geared to a counter or
recorder in such a way that the output is in units of speed.  These instruments
cannot pro-vide an offset.  Lockhart (1977) shows that some cup designs,
specifically the Meteorology Research, Inc. Model 1074 used on the Mechanical
Weather Station, can be nearly as accurately described without an offset (a=*0)
as with one (a=*O.Q3).  Table 4.2.2.4 lists the NBS data and the linear
regressions to sup-port this fact.  The residual errors from each model are
plotted in Figure 4.2.2.7.

-------
                                                             Section No.  4.2,2
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:   10   of 32
            Table 4.2.2.4  NBS Test Data for MR! Model 1074

Test
Na.

1
2
3
4
5
4
7
a
9
10
a
12
13
14
15
14
17
18
Test date

Output
Freq.
lluJ
14
28
45
m
112
240
380
500
425
755
830
1005
1255
1500
1755
•1995
2245
2530
flea 	 i
Tunnel Tunnel
Speed Speed
(tph) (1/5)
0.9 0.4
1.3 0.4
2.8 1.3
3.7 1.7
4.7 2.1
9.4 4,3
15.. 1 4.8
19.9 S.9
24.9 H.l
29.1 13.3
34.7 15.5
39.4 17.7
49.5 22.1
59.3 24.5
49.4 31.1
79.4 35.4
90,2 40.3
100.7 45.0
i - — linear Regression 	
Output
/132 r
(rps} d/sl
0.11 0.28
0.21 0.53
0.49 1.19
0.47 1.59
O.B5 2.02
1.82 4.29
2.S8 4,77
3.79 8.90
4.73 11.12
5,72 13.43
4.47 15.44
7.41 17.84
9.51 22.30
11.34 24.44
13.30 31.17
15.11 35.42
17.14 40.21
19.17 44.91
- I1/1B/75 HR1, ftltadena, Cilif.

I'-Y
(1/5)
-0.12
-0.05
-0,07
-0.04
-0.08
-0.00
8.02
0,01
-0.01
0.10
0.13
0.14
0.17
0.13
0.05
-0.14
-0.11
-0.10
Model 1074
r
X" "-Y Regression Output.- Na. 1-1S
d/si it/s) Constant 0.03201!
0.25 HU5 Std Err of Y Est 0.104914
0,50 -S.Oi R Squared 0.99994!
1.15 -3,10 Mo. of Observations is
1.54 -0.09 Degrees of Freeaot 16
1.99 -0.11
4.24 -0.03 I Coefficients 2.34145-
4,75 -3.00 Std Irr of Cotf. 0.004174
i.Bi -5.02
11.10 -0.03 v
13.41 0.09 Regression flutaut: Na. 1-1S
15.43 3.12 Constant 5
17.35 0.15 itd trr of Y Est 0.104031
22.29 fi.14 R Squared 0,9?9f4e
24.44 0.13 Do, of Observations 18
31.17 0.05 Decrees of Freedot 17
35.43 -0.15
40.23 -0.10 I CoefficienUi) 2.344297
44.93 -v.08 Std Err of Caef. 0.002744
1 132/rev.
Linear Regression of NBS Data

















s*~^
in
~^.
£
"v-x
L.
o
l_
k.





Of
0.4 -
0.3 -
0.2 -
0.1 -


-0,1 -

-0.2 -
-0.3 -
-0.4 -
— n «? -


/+
K
4-




+






Residual
^

Error Analysis
~~^\





\
\
\ ^*
\, —










^

KEY
X>















                  0          10         20         30

                                     Wind Speed (m/s)
Figure 4.2.2.7  Residual  Errors from HRI Model 1074
4-0
           5O

-------
                                                             Section No. 4,2-2
                                                             Revision No.    0
                                                             Date:    17 Sep S9
                                                             Page:    II  of 32


                     A similar analysis for a propeller anemometer is shown In
Table 4.2,2.5,   These data come from a test in the Atmospheric Environment
Service (AES) of Canada wind tunnel on a propeller anemometer being used in a
"round robin" experiment to estimate the accuracy of wind tunnel calibrations.
Each test was run for 100 seconds.  The tunnel speed is an average of one
second samples taken every ten seconds by AES,  The sensor count is a total for
100 seconds from the light chopper delivering 10 pulses per revolution.  Each
test was replicated and tests 3 and 4 were also replicated in tests 19 and 20.
Two linear regressions were run. The first, and best fit, allowed the intercept
of the X axis,  or zero offset in ASTM language, to be calculated.  The second
forced the straight line through the origin.  This latter method yields a
constant slope or pitch (meters per revolution) which when multiplied by the
rate of rotation (revolutions per second) results in wind speed (meters per
second).  The residualerror from these two regressions are plotted in Figure
4.2.2.8.

                     It is characteristic for helicoid propellers to show a
better correlation with wind tunnel speeds than does a cup anemometer. This is
because propellers generate torque uniformly without sensitivity of position.
Three-cup assemblies, on the other hand, produce three peaks and three valleys
in torque for each revolution (Lockhart, 1985).  Either type of anemo«eter can
be calibrated to an accuracy sufficient for most applications.

                     4.2.2.1.2.3  Application of accuracy specifications

                     An accuracy specification should include enough
information to define the type of accuracy intended and the method by which
accuracy claims may be tested.  Here are a few examples of accuracy
requirements.
                     In the Ambient Monitoring Guidelines for Prevention of
Significant Deterioration (PSD) (EPA, 1987a), it states that for horizontal
wind systems "Wind speed systems should be accurate above the starting
threshold to within 0.25 m/s at speeds equal to or less than 5 m/s.  At higher
speeds, the error should not exceed 5 percent of the observed speed (maximum
error not to exceed 2.5 m/s)."  In the On-Site guide (EPA, 1987b) in 8.1.1 it
states "Accuracy (error)(1)(2)             s(0.2 m/s +5H of observed)
         (1) as determined by wind tunnel tests conducted on production
             samples in accordance with ASTM D-22,11 test methods,21 (sic)
         (2) aerodynamic shape  (cup or propeller) with permanent serial
              number to be accompanied by test report, traceable to NBS,
              showing rate of rotation vs. wind speed at 10 speeds."
By implication, the latter specification refers to accuracy type A(2)» although
the expectation is that A(l) will be included by careful siting.  This
expectation must be addressed with experienced subjective judgment.

                     Assume a system is to be used in accordance with EPA
(1987b), and an "off the shelf" anemometer is purchased.  The manufacturer
states that the sensor delivers 30 pulses per revolution (ppr) with a transfer
function from revolutions per second, R (rps), to wind speed, U  (m/s), of

         U ,„., . 0.224 *   **£   . 0.244 *

                 - 0.224 (m/s) * 1.410  (m/r) R (rps).

-------
                                                        Section No.  4.2.2
                                                        Revision No.    0
                                                        Date:   17 Sep 89
                                                        Page;   12  of 32
              Table 4.2.2.5  HBS Test Data for Young 27106 at A£S
	 fiES 	 	
Test
to.

J
2
7
O
4
5
6
7
8
9
10
11
12
13
14
15
14
17
IB
If
20
Output
Count
(*)
3440
3475
10889
109ii
18108
18207
254 IS
25433
33006
33003
40143
40139
47634
47577
55102
55105
42413
42785
10930
10931
Tunriil
Speed
(B/5)
1,14
1.14
3.27
3.27
5.42
5.42
7,59
7.41
9,82
9. 63
11.96
11.95
14.13
14.14
14.35
14.30
18.53
18.54
3.29
3.28
Y¥ _____

Tunnel Output
Speed /IOOO I1
(i/s) 
-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:    13  of 32
                     Assume that an NBS test was -conducted after a year of
operation and the results provided a least squares analysis of
         U (m/s) = 0.301 (m/s) + 1.387 (m/r) R (rps).
What action is suggested by this finding?  Throughout the year the operator had
the electronics trimmed to output 0.224 m/s when the cups were not turning and
14.325 m/s when the cups were turning at 10 rps.   The A(3) error in converting
R to U is 0.00. The A (2) error can be expressed as follows:
            Generic transfer function:  U  =  0.224 + 1.410 R
                  Wind tunnel (truth):  U' =  0.301 + 1.387 R
     The error (E, m/s) is U - U' or    E  « -0.077 +• 0.023 R
     In terms of the measured speed, U, E  = -0.081 + 0.0163 U
Table 4.2.2.6 compares this error with the specification at different rates of
rotation.  The 1.6 percent over-estimation of speed by the generic transfer
function is not large enough to bother with data correction.  The data meet the
accuracy guidelines with two thirds of the allowable error unused.  At the next
calibration the system should be adjusted to the wind tunnel derived transfer
function.
                        Table  4.2.2.6  Wind  Speed Errors
R
(rps.)
0.000
1.225
3.388
6.993
14.203
U
(m/s)
0.224
1.951
5.001
10.084
20.250
U'
(m/s)
0.301
2.000
5.000
10.000
20.001
E
(m/s)
-0.077
-0,049
0.001
0.084
0.249
Allowed
(«/s)
±0.20
±0.30
±0.37
±0.45
±0.81
Used
(%)
38
16
0
19
31
                     4.2.2.1.2.4  Precision

                     The definition of accuracy describes a bias term and a
variable term akin to precision.  Traditionally, precision describes the
uncertainty with which a measuring process or instrument realizes the measured
value when that being measured  is the same thing and is repeatedly measured.
The key to finding the variability of a measuring process is to use a
non-varying subject.  In meteorology, and particularly in anemometry, it is not
possible to have, with certainty, a non-varying subject.  The ASTM subcommittee
D-22.11 dealt with this problem by writing the Standard Practice for
DETERMINING THE OPERATIONAL COMPARABILITY OF METEOROLOGICAL MEASUREMENTS -
D4430-84 (ASTM, 1984).  This work was patterned after Hoehne (1973) in which he
defines Functional Precision as the root-mean-square of a progression of
samples of the difference between simultaneous measurements made by identical
instruments collocated in the atmosphere.  Operational comparability applies to
two different kinds of instruments rather than identical ones.  This Method
recognizes that, from an operational perspective, the precision of a
measurement can be estimated by know-ing how well identical or similar
instruments measure the "same"  flow.

                     An EPA project collected data in Boulder, Colorado in 1982
to add to the literature some estimates of comparability. Finkelstein et al.
(1986) published in the refereed literature the material published by NOAA in
Raimal et al. (1984).  Lockhart (1988) re-analyzed these data and concluded

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 85
                                                             Page:    14  of 32

that the operational precision of anemometers at 10 m is no larger than 0.2
m/s. Operational precision is the standard deviation of a series of difference
measurements which is equivalent to the operational comparability with all the
bias (mostly calibration error) removed.

                     An expression of accuracy for an anemometer operating on a
10 m tower in the atmosphere can be expressed as some function of speed  [f(u)L
which comes from the wind tunnel test, A(2), plus or minus 0,2 m/s.  This
estimate does include the Influence of turbulence on the sensor since the 0.2
m/s comes from collocated cups and vane oriented propellers operating in a
turbulent summer environment.

                4.2.2.1.3  Distance Constant

                     4.2.2.1.3.1  Definition of distance constant

                     ASTM (1985a) defines distance constant as the distance the
air flows past a rotating anemometer during the time It takes the cup wheel or
propeller to reach (1-1/e) or 63 percent of the equilibrium speed after a step
change in wind speed.  The step change is specified as one which increases
instantaneously from 0 to the equilibrium speed.  The step change is simulated
by releasing a restrained anemometer in a wind tunnel running at the
equilibrium speed.  Several authors, among them Acheson (1988), Hayahsi  (1987),
Lockhart (1987),  and Snow et al. (1988), have commented on the difference
between the distance constant to an Increasing step function and the distance
constant to a decreasing step function.  The difference is larger with larger
and heavier cup wheels, as is the size of the resulting overspeeding error.
Snow et al. (1988) point out that a system including a sensor and an analog
signal conditioner will have a combination distance and time constant.

                     4.2.2.1.3.2  Measurement of distance constant

                     Most manufacturers will provide the distance constant of
their product.  These are usually derived from tests of prototype sensors
during the development phase of the product.  The variation froa individual to
individual in a production model is not large nor important.  It is Important
to use a standard test and standard definitions if distance constant
specifications are to be meaningfully compared to other designs and
requirements.

                     EPA (1987a) does not specify a distance constant for
anemometers.  EPA (1987b) does suggest  in the Instrument Procurement section
8.1 a distance constant of <5 m at 1.2 kg/m (standard sea-level density).  As
with accuracy, this reference uses a footnote to specify the ASTM test method.

                     The reason why distance constant is included is to  urge
users to buy high quality responsive sensors.  Heavy sensors with long distance
constants are more likely to produce overspeeding errors, which overstate  the
average wind speed.   If they are used to measure turbulence, they will
underestimate sigma u because of a failure to respond properly to eddy sizes
smaller than twice the distance constant.

-------
                                                              Section No.  4.2.2
                                                              Revision  No.     0
                                                              Date:   17 Sep 89
                                                              Page:   15  of 32
                4.2.2.1.4  Off-Axis Response
                 This  specification,  while  included  in ASTM  (1985)  and
recognized  in  the  literature as a source of error,  is not included in EPA
requirements or  suggestions.   It is  mentioned here  for completeness and in
anticipation of  future specifications when more data have been published on the
subject.

                 The off-axis errors  from helicoid propellers are nearly cosine
errors.  When  a  vane-oriented propeller is turned in a wind tunnel so that the
wind is at  some  angle to the axis of rotation of the propeller,  the propeller
slows down.  The indicated speed from this misorientation of the propeller is
nearly equal to  the total speed tines the  cosine of the  angle of
tnisorlentation.   That is, if the indicated speed from a  propeller  is 5.00 m/s
and the propeller  is  being held 10 degrees off the  true  axis of the flow by the
aligning wind  vane, the true speed is the  indicated speed  (5) divided by the
cosine of 10 degrees  (0.9848) or 5.08 m/s.   In natural turbulent flow, a vane
located behind the propeller jiay not keep  the propeller  perfectly  aligned with
the wind.   Small misalignments result in small errors since the cosine of a
small angle is nearly one.

                 The off-axis errors  from a cup anemometer with a properly
oriented vertical  axis will depend on the  design of the  cup wheel  and the angle
from horizontal  from  which the wind  reaches  the cups.  MacCready (1966) and
Kondo et al. (1971) show that cup anemometers overstate  the wind speed when the
air flow is not  horizontal.   Kondo shows the overestimation. by the cups tested
to be 5 percent  when  the standard deviation  of the  elevation angle is 17
degrees and 10 percent at 25 degrees.  Siting on ridges  or  building tops or
anywhere the distortion of the flow  over an  object  produces a steady
non-horizontal flow will result in errors  which will be  unknown.

                 The figure from MacCready  (1966) showing the response of
various anemometers to the elevation angle of  the wind  is  reproduced here as
Figure 4.2.2.9.
 -*0  _50  -40  -30   -20  -10   0   10   ZO   30   40   SO  60

              8 ELEVATION ANCLE (DEGREES)
                                                                •(I) coi 6 CURVE

                                                                m earn3 « CURVE

                                                                (3) STANDARD SMALL CUPS

                                                                (41 SENSITIVE SMALL OJPS

                                                                (St TOTAL SPEED SENSOR

                                                                (41 YELOCrrrVANE
                                                       .—	~	(7) YECTORVANE
                                                         (AXIS CONSTRAINED HORIZONTAL!
Figure 4.2,2.9  Anemometer response to off-axis flow

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   16  of 32
                Consider the fact  that a wind of  5 m/s with an elevation angle
of 30 degrees will have a horizontal component of 5 x cos  (30) - 4.33 m/s.  If
the presumption is that the cup anemometer  is providing the speed of the
horizontal component of the wind,  and If the cup performs  like a "total speed
sensor" in the range of ±50 degrees as the  figure suggests, the 5 m/s the cup
reports is a 15% overestimatlon of the true horizontal speed of 4.33 m/s,  A
propeller anemometer will report the horizontal component  because it does have
a nearly-cosine response.  Operating side by side in this  30 degree wind, the
cup will report 5 m/s and the propeller will report 4.33 m/s and each will be
"right."

                In addition to the horizontal component dilemma, the cup
anemometer tends to overestimate even the total wind.  This is particularly
noticeable when the air is rising  and flows past the support column creating a
wake which interferes with the normal cup aerodynamics.  The figure shows this
effect to be about 10% at +30 degrees for "standard small  cups."  This 30
degree rising air example suggests that the side fay side anemometers mentioned
above will really be reporting 5.5 m/s (10% off-axis error for the cup) and
4.33 m/s for the true horizontal speed from the propeller, or a 27%
overestimatlon of the horizontal component  by the cup anemometer.

-------
                                                             Section No, 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   17  of 32


       4,2.2.2  Wind Direction

                4.2.2.2.1  Threshold

                     4.2.2.2.1.1  Definition

                     As with wind speed measurement, a key to a good wind vane
for air pollution applications is a low threshold.  The threshold is the one
performance characteristic which will certainly change with time because of
bearing degradation.  Most wind vanes use potentiometers to convert position to
output voltage.  Potentiometers have bearings or bushings which will wear and
add to the starting threshold.  There is not a standard definition for wind
vane threshold, although the ASTM Standard Test Method for DETERMINING THE
DYNAMIC PERFORMANCE OF A WIND VANE (ASTM, 1985b) offers the following
candidate.  The 6  in the definition is the equilibrium direction of the vane
                 3
in a wind tunnel at about 10 m/s.

          Starting threshold  (S , m/s) is the lowest speed at

          which a vane will turn to within 5  of fl  from an

          Initial displacement of 10°.

Even this definition runs into some problems in interpretation.  If the vane
must move at least from 10  to 5  at^the threshold speed, Is the offset
sensitivity really 5° rather than 10 ?

                     The requirement in EPA (1987a) for PSD applications states
"Wind direction and wind speed systems should exhibit a starting threshold of
less than 0.5 meter per second (m/s) wind speed (at 10 degrees deflection for
direction vanes)."  Does this mean that a vane that moves from a 10°
displacement to 9.5  at 0.5 m/s has a starting threshold of 0.5 m/s?  The newer
EPA (1987b) on-site guidance says

          "Threshold (1)               s 0.5 m/s
          (1) as determined by wind tunnel tests conducted on production
          samples in accordance with ASTM D-22.11 test methods."
                                                               o     o
The reason the ASTM committee required the vane to move from 10  to 5  was to
relate the starting threshold to accuracy.  With wind speed, there is a way to
correct for threshold nonlinearity; for wind direction there is not.  It seemed
best to establish the range of operating speeds to correspond to the range
where accuracy requirements are met.  ASTM assumed 5  for wind direction as a
reasonable accuracy.

                     When torque measurements began their use as a measure of
starting threshold, the question became clearer.  If the vane is required to
move to 5  there should be enough torque developed by the wind speed working oa
the tail area exposed at 5  from the wind tunnel centerllne or the true wind
direction to turn the shaft assembly and transducer.  This sounds like a 5
threshold requirement, and perhaps that Is a better description.  As will be
shown later, there is a big difference between the torque developed at some
speed at 10  and the torque developed at the same speed at 5°.  The nature of a
standard test method is less  important than the application of a standard
method everyone uses and regulatory performance requirements consistent with
that test method.  This handbook will use the 10  offset moving to 5° on
release in the wind tunnel (the ASTM method) as the criteria for starting
threshold.  The relevant torque for this definition is that at 5 .

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   IS  of 32
                     4.2.2.2.1.2 Threshold measurement
                     The measurement of starting threshold requires a wind
tunnel capable of accurate operation below 1 m/s.  One standard methodology is
defined in ASTM (1985b) and described in Finkelstein (1981).  Just as it is
with wind speed, it Is possible to measure the torque which results from the
force of the wind on a wind vane as the torque measurement device holds the
vane at some angle from the wind tunnel centerline, say 10°.  Lockhart {1978}
describes wind tunnel test data using two very different wind vane designs, the
front-damped Meteorology Research, Inc. (MRI) Model 1074 and the more
traditional Teledyne Geoteeh (TG) Model 53.2.

                     Another body of wind vane torque data exists as a result
of tests run by the R. M. Young Company (RMY).  Their tests used a DC "torque
motor" as the transducer for vanes mounted in their wind tunnel.  The torque
motor current was linearly correlated to torque measured with a series of
Waters Torque Watches.  The torque motor drove the vane to each of four
positions, plus and minus five degrees and plus and minus ten degrees from the
wind tunnel centerline.  A measurement of current was taken at each position
and at each of 12 wind speeds varying from 0.3 to 6 m/s, depending on the vane
design. Table 4.2.2.7 lists the average constant, K, which was found by a
linear regression of the motor current (torque) to the square of the wind
speed, with the intercept forced to zero.  They tested all of their products
along with some vanes from other manufacturers.

          Table 4.2.2.7 - K Values for Vanes at Two Angles to the Hind
Vane Type
Offset Angle
5°
K
Wind Sentry (RMY 033015
Wind Monitor (RMY 05103)
Wind Monitor AQ (RMY 05305)
Propvane (RMY 08003)
Microvane (RMY 12302)
Bivane-19 cm fin (RMY 17003)
Anemometer Bivane (RMY 21003)
Propeller Vane-23 cm (RMY 35003)
Long Vane (Vaisala WAV 15)
Short Vane (Vaisala WAV 15)
Black Aluminum (Met One 024A)
High Damping Ratio (Met One024A)
F460 Vane (Cllmatronics 100075)
where: K = TAJ2, E * Std. Err.
1
10
16
15
25
14
17
19
3
2
.8
.6
.8
.9
.0
.5
.1
.0
.6
.0
13.8
19
16
of
.9
.0
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
Coeff.
E
006
080
126
061
414
188
141
127
047
015
181
194
322
and
e
10
o
K
3.
23.
37.
38.
57.
37.
45.
46.
7.
4.
28.

29.
r B
7
3
0
8
5
6
6
5
8
3
4

8
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.

0.

E
017
114
260
304
760
367
457
378
049
029
394

497


r
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.

1.
1
2
2
4
3
6
7
4
2
2
1

9
K(10°)/K(5°)
                     The  ratio  of the  10 degree  K value to the 5 degree K value
 seems  to be  lower for  high aspect ratio vanes.   A square vane has an aspect
 ratio  of 1.  The Propvane,  Microvane,  Bivane  and Anemometer Bivane are examples
 of designs with an aspect ratio of 1.  A rectangular vane which is two times  as
 high as it is  long  (along the tail boom) would have an aspect ratio of 2.  The
 Wind Monitors  are examples of this design.  The  F460 vane has an aspect ratio

-------
                                                            Section No. 4.2.2
                                                            Revision No,    0
                                                            Date:   17 Sep 89
                                                            Page:   19  of 32
 of 4 and a .K.ratio of 1.9.   The "high aspect ratio" TG Model S3.2, whose  torque
 data (natural log of torque vs.  natural  log of displacement angle) are shown in
 Figure  4.2.2.10, has an aspect ratio  of  6 and has "a K value ratio of 1.4  at
 0.45 m/s.  Differences in torque between 5 degrees and 10  degrees could not be
 measured at 2.2 m/s (the K value ratio therefore equals one at  that speed).
 The High Damping Ratio (Met One) also has a high aspect ratio and also could
 not provide a stable torque reading at 10 degrees.    The  MRI 1074 (aspect
 ratio of 2) has a K ratio of 2.6.   This  design Is more difficult to compare to
 other vanes because of its front damping vane.

        Wind  Vane  Torque  Data -  Two  Vanes
                      1975 Data  - (see Lockhart 1978)
  o
     5 -
     4 -
  0)
  D
  tr

  O  .3

  c
              KEY:
            MR! Model 1074
            TG Model 53.2
     2 -
                                                                      -500
                                                                      -100
                                                                      -so
E
o
i
a>

v
3
C7
O
                     In Displacement Angle  (deg.)
Figure 4.2.2.10  Torque measurements as a. function of vane angle.

                If  the starting torque of the shaft of a direction vane bearing
and transducer assembly is  to be  interpreted in terms of wind speed, an
expression of torque as a function of speed is required.   Each expression is
specific to the vane design and an offset angle.  Take, for example, the Wind
Monitor AQ shown in Table 4.2.2.7.  The expression for a 10 degree offset is

                              T  - 37 U?

                If  a starting torque were found to be 5.9  g-cm,  that
measurement can be  expressed as a threshold wind speed of  0.4 m/s (0.9 nph).  A
0.41 m/s wind at a  10 degree angle from the vane position  will produce enough
torque to move the  vane closer to the wind direction.  The expression for this
wind vane for a 5 degree offset is

                              T  - 16.8 U?

The same starting torque of 5.9 g-cm will require a wind speed of 0.6 m/s (1.3
mph) to move the vane closer than 5 degrees to the true wind direction.

-------
                                                             Section No.  4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:    20  of 32
           Table 4.2.2.8 - Wind Vane Torque vs.  Wind Speed and Angle
Wind
Speed
U
(m/s)
1.3
2.2
4.5
Offset Angle 0 (deg. )
o
5 A
T T
(g-cm)(g~em)

72 69
216 217
<°\
T T
(g-cm) (g-cm)
72 76
166 177
562 557
»\
T T3
(g-cmHg-cm)
108 97
324 278
1152 1163
20\
T T
(g-cm) (g-cm)
144 ISO
432 430

T is the measured torque holding the vane at offset angle 9
K is the linear regression coefficient when a * 0
A
T is the predicted torque using K. from the following;
A 1 fi A 16
Tj- 19.55 U T2= 50.22 U
A 2. A 2
T = 57,47 LT 1 » 88.81 IT
••J TE
                     The data from Lockhart (1978) for the MR! Model 1074 are
shown in Table 4.2.2.8 to demonstrate the complexity of the dynamic performance
of some vane designs.  A simple expression Is useful to convert a torque
measurement to a wind speed.  The simple vane designs listed in Table 4.2.2.7
fit a U  expression quite well.  The
1074
degree and 10 degree data for the Model
on the log-log plot of Figure 4.2.2.11.
      define a different slope than U
An expression of U1"6  fits  the data well enough  to use to extrapolate  the
experimental data for this vane design to other wind speeds.  The physical
reason for this unusual dynamic performance is probably related to the effect
of the front damping vane and the relatively large support column.  The
vortices shed by the column only effect the rear vane.

                     The question remains, should the 5 degree K value be used
or the 10 degree K value?  For the purpose of making a conservative estimate of
starting threshold for performance accurate to 5 degrees,  the 5 degree K value
is recommended.  The user should not expect this torque-defined threshold to
agree with the "starting threshold" published by manufacturers.  Only after a
test is specified, like the ASTM test, can a 5 degree K value be expected to
agree with the data sheet values.

                     4.2.2.2.1.3 Torque measurement

                     Starting torque measurements of a wind vane may be  made in
either of two general ways.  If the vane can be removed, a torque watch  can be
used to measure the starting torque of the bearing assembly and transducer (see
Figure 4.2.2.3).  For this method to be most accurate,  an equivalent weight of
the removed vane must be placed on the shaft to simulate the end loads of the
shaft of the bearings.

-------
                                                          Section No. 4,2.2
                                                          Revision No.    0
                                                          Date:   17 Sep 89
                                                          Page:   21  of 32
          Wind  Vane  Torque  Data  - MRI  1074
                      1975 Data - (see  Lockhart, '978)
          1.3 m/s     v 20°
                                             T = 19.55 U1'6
                                                                       •1,00"
                                                                           E
                                                                           o
                                                                           V
                                                                           £•
                                                                           u.
                                                                           o
                                                                        100
                                                                      - 50
                0.4
1.2
1.4
Figure 4.2.2.11
         0.6       0.8       1
          In Wind Speed (m/s)
Torque measurements as a function  of wind speed.
1.6
                    If the vane cannot be removed or  the  choice is to not
remove it,  the starting torque can be measured by imposing a force at a
measured radial distance from the axis of rotation, A spring-type gram scale
at 10 cm from the axis of rotation will yield g-crn after dividing by 1O.  On
some designs it is impossible to impose the force at 1 cm.  In the interest of
accuracy,  it is better to use a longer distance so the length part of the
measurement can easily be just a few percent. Of course the trade-off for
accurate distance is small force, an equally troublesome source of uncertainty.
Figure 4.2.2.12 shows different gran scales used on a  Young Wind Monitor AQ.
Figure 4.2.2.12  Starting torque measurements on & wind vane.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   22  of 32

                      If the vane Is left on, the space used for the measurement
must be devoid of any air movement.  Human breath provides a force which can
bias the measurement.  It is also important that the axis of rotation be
vertical to negate any imbalance in the vane assembly.

                      For either method, the full 360 degrees of rotation of the
vane should be challenged with the highest torque found being reported as the
starting torque  (worst case).

                4.2.2.2.2 Accuracy

                      4.2.2.2.2.1 Accuracy definition

                      There is no transfer function for a wind vane comparable
to that for an anemometer, unless the conversion of shaft position to output
voltage is taken to be such a function.  The vane is assumed to be accurately
placed, on average, downwind from the axis of rotation, when the wind is steady
and its speed is well above the threshold.  If the vane is bent In some way, a
bias will be introduced (see Dynamic Vane Bias later in this section).  This is
seldom large enough to be of concern.

                      The accuracy of the sensor is described by how well the
shaft position is reported by the transducer and signal conditioning circuit.
The accuracy of wind  direction must include the accuracy with which the sensor
is sited with respect to TRUE NORTH.  Any error in orientation will be a true
bias and can be removed from the data at any time the facts become known.  The
"facts" in this case  mean a rigorous quality control program which requires a.
site log to indicate  any servicing of the sensors.  The "true bias" can change
if the sensor is removed and reinstalled without "as-found and as-left"
orientation measurements in the log book.  Any possible undocumented change can
negate data correction for orientation.

                      The requirements for accuracy include EPA (1987a) which
states "Wind direction system errors should not exceed 5 degrees, including
sensor orientation errors."  In EPA (1987b) it says

    "Accuracy (errorHi)      *3 degrees relative to the sensor
                               mount or index (s5 degrees
                               absolute error for installed system)

     (1) as determined by wind tunnel tests conducted on production
     samples in accordance with ASTM D-22.11 test methods."

The footnote is in error.   There is nothing in the wind tunnel test which
relates to wind direction accuracy.

                      4.2.2.2.2.2 Measurement of sensor accuracy

                      The simple procedure for this measurement requires some
fixture which provides for steps in the direction vane shaft position of known
size.  There are innumerable devices and methods for this procedure, many of
which will be described in the calibration section (4.2.5.2).  One device which
can move the shaft in 60 degree increments is shown in Figures 4.2.2.13.  The
important criteria are stability and knowing that the error band for the
fixture is on the order of 0.1 degrees of arc.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   23  of 32
Figure 4.2.2.13  Wind direction calibration fixtures from Teledyne
                 Geotech (left) and Met. Standards Institute (right).

                     Typically, potentiometers used for wind direction will
have a linearity of about 0.5 percent, 1.8 degrees In 360 degrees.  A table of
angles and output values will usually fall within a range between -1.8 degrees
and +1.8 degrees when the bias is removed by subtracting the average error from
each error.  This statement  is true when the open sector of the potentiometer
is ignored (for 360 degree mechanical and about 352 degree electrical systems,
see Section 8,2) or when errors In 540 degree format switching systems are not
considered. Other contributions to sensor error, such as hysteresis,
out-of-round and signal conditioning errors, when added to the linearity error
mentioned above should provide an error band not larger than -3 to +3 degrees
relative, or 6 degrees if the bias has not been removed.

                     An example of audit data shown in Table 4.2.2.9 describes
the performance of one wind  vane when challenged with a 60 degree fixture.  The
fixture settings and the displayed digital output ofthe system are listed.  The
system had a 540 degree range and a 5 volt full scale output.   The output is
converted to nominal voltage to show how the 540 range works.   (Degrees per
volt = 540 / 5 = 108)
                                              O
                     The average error of -3,4  was calculated without using
the obvious "open section" values marked by "*."  When the fixture is installed
the vane substitute is set in the 180  location and then rotated until the
output is about 180.  This need not be precise since the average error provides
a means of normalizing the data by removing the initial bias of approximate
setting.  The linearity of the potentiometer-signal conditioner can be seen in
Figure 4.2.2.14.  Except for the "open sector" near 360 , the error is within a
± 3  band, including the 540  format switching error of about 1°.

-------
                                                      Section No. 4.2.2
                                                      Revision No.    0
                                                      Date:   17 Sep 89
                                                      Page:   24  of 32
      Table 4.2.2.9 - Relative Wind Direction vs.  Output Direction
Fixture
Setting
A
(deg. )
180 ccw
120 ccw
060 ccw
360 ccw
300 ccw
240 ccw
180 ccw
120 ccw
180 cw
240 cw
300 cw
360 cw
060 cw
120 cw
180 cw
240 cw

cw
System
Output
B
(deg. )
177
114
054
001
298
238
176
115
177
239
292
001
056
115
177
239
average
Nominal
Voltage
B/108
(volts)
1.639
1.056
O.SOO
0.009
2.759
2.204
1.630
1.065
1.639
2.213
2.704
3.343
3.852
4.398
4.972
2.222
error C
is clockwise, ccw
Error

E=B-A
(deg. )
-3.0
-6.0
-6.0
1,0*
-2.0
-2.0
-4.0
-5.0
-3.0
-1.0
-2.0
1.0*
-4.0
-5.0
-3.0
-1.0
= -3.4 (*
Normalized
Error
E-C
(deg. )
0.4
-2.6
-2.6
4.4«
1.4
1.4
-0.6
-1.6
0.4
2.4
1.4
4.4*
-0.6
-1.6
0.4
2.4
values
Output
B-C
(deg. )
180.4
117.4
057.4
005.4
301.4
241.4
179.4
118.4
180.4
242.4
301.4
005.4
059.4
118.4
180.4
242.4
excluded)
is counterclockwise
          Relative Wind  Direction Accuracy
             Actual 16—Point Audit Results - 540  Degree Format
3 -
4 -
, — ^*.
cr» 3 -
0)
"O
k_ 2 -
O
L.
L*-1 -j _
X!
O>
1X1
O
E
0 -1 -
"Z.

-2 -

Number indicates point sequence
1—8 counterclockwise
8—16 clockwise










'\,
"X--,,^ -
^^











*• 13
•vv
%""-> :

K«.
+ 3
1








:
•
P*
/ j
/
' p:'i 4
•''
^2 •




A
t
/




I: 10,16
^
\





^""-^
/ +6 \ *t>,11
/' *"-i.
^^
: |/g •] ^
"/' ' '

h 7













~~~*^
*"».
"'•-.s






4,12

















l ( l I
060
300
                        120      180      240
                       Fixture Position (deg.)

Figure 4.2.2.14  Results of a wind vane audit using 60° steps.
360

-------
                                                             Section No. 4.2,2
                                                             Revision No.    1
                                                             Date: -   17 Sep S?
                                                             Page:    25  of 32


                     An analysis of this type helps to optimize the accuracy cf
the orientation.   If an orientation target is at 120  TRUE, when the vane is
pointing from 120  the output should read about 118 .   This effectively centers
the error band (see 4.2.4.3.2 on orientation).

                     4.2.2.2.2.3 Measurement of orientation accuracy

                     Orientation error is an important part of the measurement
error, but it cannot be considered until the sensor is installed in the field.
The accuracy of the orientation includes the accuracy in finding TRUE NORTH arc
the accuracy with which the vane is aligned to TRUE NORTH.  Use different
methods for finding TRUE NORTH.  Methodology for orientation is given in
4.2.4.3.2.

                     4.2.2.2.2.4 Expression of Accuracy

                     An accuracy specification should include enough
information to define the type of accuracy intended and the method by which
accuracy claims may be tested.  There is no requirement for traceability to N3S
for wind direction.   The measurement of relative direction is a fundamental
division of a circle.  The measurement system can be bench tested by basic
methods yielding a clear expression of the errors associated with a position
angle vs. system output transfer function.  The errors are mostly conditional
biases which are small enough  (less than one percent of 360 degrees) to ignore.

                     The orientation error is a pure bias which cannot be bench
tested.  The total error, a simple sum of the two parts (root-sum- square, RSS.
combination is only legitimate with random errors, not biases), can only be
found after installation.

                     4.2.2.2.2.5  Precision

                     The definition of accuracy describes a bias term and a
variable term akin to precision.  A comparability test (see 4.2.2.1.2.4) will
show that two vanes properly sited and perfectly oriented will report the 20
minute scalar average directions with a difference of less than two degrees,
i.e. precision is ± 2 .

                     An expression of accuracy for a wind vane operating on a
10 m tower in the atmosphere can be expressed as the relative accuracy plus
orientation accuracy and ± 2° for precision.  For a collocated test CLockhart.
1988), the orientation error can be estimated by the average difference between
the subject wind vane and a collocated wind vane perfectly oriented.  If the
orientation error is found to be large, and if a quality control system has
provided records of maintenance showing the orientation has not been changed, a
bias correction can be applied^  The accuracy of the data corrected for bias is
then the relative accuracy ± 2  .

                4.2.2.2.3  Delay Distance (Distance Constant)

                     4.2.2.2,3.1  Definition of delay distance

                     ASTM (1985b) defines delay distance (D) as the distance
the air flows past a wind vane during the time it takes the vane to return to
50 percent of the initial displacement.  The value for this sensor
specification is found in wind tunnel tests, as described in Finkelstein
(1981).  The initial displacement is 10 degrees and D is the average of a
series of tests at 5 m/s and 10 m/s using displacements on both sides of the
tunnel centerline.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   26  of 32

                     The specification in EPA (1987a, PSD) says "...the
distance constant should not exceed 5m."  In EPA (1987b, On-Site) the
specification says  '      •           '

     "Delay Distance (1)                 35 m at 1.2 kg/m3 (standard
                                          sea-level density)
      (13 as determined by wind tunnel tests conducted on production
         samples in accordance with ASTM D-22.11 test methods."

                     4.2.2.2.3.2  Measurement of delay distance

                     Measurement requires a wind tunnel of reasonable size and
quality.  The width of the tunnel should be at least three quarters of the
overall length of the wind vane to be tested.  With the small displacement
angle of 10 degrees (about 3 percent of full scale), it is hard to conduct this
test in the open atmosphere.

                     This specification is strictly a sensor dynamic
performance specification.  Any time constants In the signal conditioning
circuits will dampen the apparent sensor response and make D larger than it is
for the sensor.  One could argue that It is ONLY the combination of D and the
time constant of the signal conditioner that should be considered in meeting
the regulatory requirements for performance. The 5 m maximum for D is roughly
equivalent to a time constant of 0.5 seconds at 10 m/s wind speed and aim
vane.  For this and other reasons it is best to keep the time constant of the
signal conditioning circuits to 100 ms or less.  For the same reason, it Is
necessary to use high speed recording equipment for the wind tunnel tests.  At
10 m/s, aim vane reaches the 50 percent D value in 100 ms.  If one wants
resolution to find D to 10 percent of the true value (0.1 m), a 5 ms resolution
in the data is desirable.

                4.2.2.2.4  Overshoot or Damping Ratio

                     4.2.2.2.4.1  Definition of overshoot or damping ratio

                     ASTM  (1985b) defines Overshoot  (Q) as the ratio of the
amplitudes of two successive deflections of a wind vane as it oscillates about
8  after release from the offset position, as expressed by the equation
                                   e
                                    n
where 9  and 9      are the amplitudes of the n and  (n+1) deflections,
       n      
-------
                                                             Section No. 4,2-2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   27  of 32


                  The specification in EPA (1987a» PSD) says "The damping
ratio of the wind vane should be between 0.4 and 0.65..."   In EPA (1987b,
On-Site) the specification says

              "Damping Ratio (1)        aO.4 at 1.2 kg/m3  or

               Overshoot (1)            <25% at 1.2 kg/m3

               (1) as determined by wind tunnel tests conducted on
               production samples in accordance with ASTM D~22.ll test
               methods."

                     The subject of dynamic wind vane performance is thoroughly
discussed in MacCready and Jex (1964), Gill (1967), Weiringa (1967) and Acheson
(1970).
                     4.2.2.2.4.2  Measurement of overshoot

                     The measurement of overshoot also requires a good wind
tunnel and sensitive, fast response recording systems.  A series of tests were
conducted by Lockhart in 1986 in pursuit of a wind vane design with a 0.6
damping ratio.  A sketch of the results of this unpublished work is shown in
Figure 4.2.2.15 to provide an example of how various vane designs compare in
overshoot and delay distance.  One of the requirements in the ASTM method is an
initial offset of 10 deg.

                     4.2.2.2.5  Dynamic Vane Bias

                     The Dynamic Vane Bias (0 ,  deg. ) is the displacement of
                                             B
the vane from the wind tunnel centerline at 5 ra/s.  This measurement will
identify wind vanes with unbalanced aerodynamic response because of damage
(bent tail) or design.  This is a screening specification not needed or used in
any application requirements.    The ASTM method measures this difference, if
any, and disqualifies the vane if the difference is greater than one degree.

       4.2.2.3  TURBULENCE

                4.2.2.3.1 Definition

                The Glossary of Meteorology (Huschke, 1970)  quotes Sutton
(1955) defining turbulence as a. state of fluid flow "in which the instantaneous
velocities exhibit irregular and apparently random fluctuations so that in
practice only statistical properties can be recognized and subjected to
analysis.  The situation is, in fact, analogous to that accepted unreservedly
in the field of molecular physics..."  The definition is ended with a quote
from the Bible"

               The wind bloweth where it  listeth and  thou hearest
               the sound thereof but canst not tell whence it cometh
               and whither it goeth: ...  John 3:8

                From the standpoint of wind measurement, then, turbulence is
not measured, it is calculated.   From the standpoint of quality assurance,
turbulence is a difficult subject to control.   It is possible to define the
measurement samples from which the statistical properties are calculated.  It
is possible to define the algorithm by which the samples are summarized.  The
relationship between the algorithm and the application or model is also
important, but it is beyond the scope of this handbook.

-------
                                                                Section No.  4.2.2
                                                                Revision No.     C
                                                                Date:   17 Sep 89
                                                                Page:   28  .of 32

t
10*
_L
t
10*
™L
1
10*
_i_
t
10*
JL

1
10*
}

1
10*


f
1.
10
	 1 	
t
10*
_1_
1
/

/

/

J

f
s
f
r


f
f
Delay Dis
s"
s^
^^
"V


\
x. ^^
1
f
f
f



J
/
J
/
f
'

r\



Si


\
V_/


^ ^^m ""^
A^y'"


M^^
tance (50% recovery)
-~~^
Thies (experimental)
697 grams
MRI 1074
656 grams
Thies
600 grams
'
Climatronics F460
^2S1 grams

* "" **m_
^~«-*^
MSI (experioental)
191 grams

Climet
170 grams



Vaisala
92 grams
Id (experimental)
72 grams
   012       5
  Distance (B) at 5 m/s                              Neteorological Standards
                                                    Institute - March 5. 1986
Figure 4.2.2.15  A  sample of  the dynamic response of some wind! vanes

-------
                                                             Section No. 4.2.2
                                                             Revision No.    C
                                                             Date:   17 Sep 89
                                                             Page:   29  of 32


                4.2.2.3.2 Direction Measurement for Sigma Theta

                The most common turbulence property routinely reported is sigss
theta, the standard deviation of a series of horizontal wind direction samples.
Most, but not all, of the following will also apply to sigma phi, the standard
deviation of a series of vertical wind direction samples.  Among the
specifications which are important to the direction measurement used to
calculate Sigma theta is delay distance which limits at the small end of the
eddy size spectrum the eddy sizes to which the vane can react.  If the vane has
a delay distance of 5 m, it will not detect energy from eddys smaller than 5 «
because the vane cannot react to them.  If 1 m eddy sizes are important to the
diffusion being estimated, use a wind vane having a delay distance of 1 m or
less.
                Another important specification is overshoot or damping ratio.
Vanes will overshoot when correcting for a direction change. If the overshoot
ratio is 0.5 (or 50%), more variability will be reported from the same
turbulent flow than is reported by a vane with an overshoot ratio of 0.25  (or
25%).  The relationship between overshoot ratio and damping ratio is given in
Table 4.2.2.10 as calculated by the equation found in 4.2.2.2.4.1.

          Table 4.2.2.10 - Overshoot Ratio vs. Danping Ratio
Over
Ratio
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05

shoot
Percent
100
90
80
70
60
50
40
35
30
25
20
15
10
5
! denotes
Damping
Ratio
0.00
0.03
0.07
0.11
0.16
0.22
0.28
0.32
0.36
0.40
0.46
0.52
0.59
0.69
PSD range
                Sigma calculations are biased by any averaging built into  the
signal conditioner.  They are also subject to error if external noise gets into
the output, a dilemma for circuit designers.  A compromise might be to filter
out any noise at frequencies higher than 20 Hz (0.5 m at 10 m/s).  In winds
above 20 m/s, this filter would degrade data from a wind vane having a delay
distance of 1 m.           Turbulence from mounting structures upwind of  the
vane will bias  the slgma value.  Out-of-balance conditions with a vane
measuring sigma phi will also bias the statistic, particularly at the low  wind
speeds.  Un-filtered noise from potentiometers will add an error to the natural

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   30  of 32

variability of the wind.   These are problems which are best detected by
inspection of strip charts or oscilloscope traces,

                4.2.2.3.3 Statistical Summaries

A few basic concepts will help In considering the specifications of the
statistical algorithm used and the representativeness of the value calculated.
Here again, careful definition will help understand what the circuits and
logical networks are doing to the input samples.  EPA (1987b) devotes 44 pages
to Meteorological Data Processing Methods.

                Representativeness is the important concept to keep in mind
when examining strange or unusual data.  Samples of wind direction taken over a
short period of time (seconds to a minute or two} are likely to exhibit nearly
normal or Gaussian distribution.  As the time gets longer (a few minutes to an
hour or more), physical dynamics driving the flow in the surface layer may
provide different shapes.  The most common of these might be the bi-modal
distribution resulting from land-water, mountain-valley, day-night or
meso-scale convective flow systems.  Whatever the driving forces, a bi-modal
distribution cannot be usefully represented by a mean and standard deviation.
This is to say that a data sampling and processing system may work perfectly
and produce numbers which have no physical meaning. From a specification
standpoint, tests for "working perfectly" are possible and should be used.

                The method used by the wind direction system to describe the
position of the vane in the series to be statistically described must be
thoroughly described and understood.  The most common error in the past,
perhaps even made today,  is to do nothing.  If the output voltage unambiguously
represents an azimuth angle, and if samples of voltage are described with the
statistical parameters of mean and standard deviation, and then expressed in
units of azimuth angle, great errors will result.  These errors are a result of
a discontinuous range of output voltage.  If 001-360 degrees are represented by
0-1 volt, samples clustered around 360 will contain some near 0 and some near
1.  The mean of 0.5 will be ISO degrees away from the mode.

                When analog ink recorders were used exclusively with 360 degree
formats, it was common to see the paper painted red by the pen going back and
forth through full scale, effectively obliterating any data.  There are several
ways to avoid or minimize the "crossover" problem.  System specification should
define how this will be done.  The most common method for minimizing this error
is to use a "540 degree"  format.  Systems were designed with dual
potentiometers or dual wipers 180 degrees out of phase.  When the wiper moved
Into the gap, circuit switching would change to the center of the other circle.
This switching would be invisible in the output at the 1/3 and 2/3 scale
points, but when the voltage went beyond full scale, it would switch to  1/3
scale and when the voltage went to zero, it would switch to 2/3 scale.  This
format completely eliminated the pen painting problem and drastically reduced
the output voltage switching, but some large pulses remained to occasionally
bias sigma calculations.

                With the advent of microprocessors and digital computers it
became possible to combine the samples without any large pulses.  One method
uses a unit vector sum to find the resultant vector direction (average
direction).  With an assumed wind speed of 1 m/s, each sample of wind direction

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   31  of 32

is converted from polar coordinates  (001-360 degrees) to cartesian coordinates
(N-S and E-W components in meters).  The components are added or subtracted over
the sample period and the resultant  vector direction is found by converting the
final coordinate sums to polar coordinates through an arc-tangent calculation.
The standard deviation of each sample about the resultant vector direction is a
straight forward process on. the differences of the samples from the "average."
Of course, this same method can be used with the true speed-weighted wind
vector samples.

                Some automatic systems currently available pick an assumed
direction, usually  the mean of the last period, take the digital difference in
degrees (limited to 180) of each sample from the assumed mean, and find the
standard deviation  and the mean difference for the period.  The standard
deviation about the mean is the same as the standard deviation about the mean
plus a constant.  The mean difference plus the assumed direction is the true
mean (limited  to 360 degrees).

                There are also other algorithms which estimate the standard
deviation (see Turner, 1986).  It  is only necessary from the standpoint of
quality assurance to know that the method used is being satisfied with the
samples taken from  the measurement system.

                The sample size is specified in EPA (1987W as 360 samples to
estimate the standard deviation to within 5 or 10K.  Lockhart  (1988) found an
apparent bias, not  a random error, when the standard deviation was estimated
from 120 samples over a 20 minute  period.

                Most models accept data representing one hour.  Sigma theta for
60 minutes is  influenced by the changing wind direction during the hour.  It is
recommended  (EPA, 1987b) that four 15-minute slgraa theta calculations be
combined to provide a "one hour" value for the purpose of selecting a
Pasquill-Gifford stability class.  The method is
A(l-hrJ
2 2 2 2

-------
                                                              Section  No.  4.2,2
                                                              Revision No.     0
                                                              Date:    17  Sep 89
                                                              Page:    32   of 32


is excluded from this value but  Included  In the true  standard deviation  of the
direction about the hourly mean.   If the  standard deviation about the hourly
mean is required for a concentration distribution analysis,  the correct
formulation for 
-------
                                                             Section No. 4.2.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  5
4.2.3  ACCEPTANCE TESTING
The procurement document, purchase order or contract, should be specific at
least in terms of required performance specifications.  "Required"'in this
context may only mean that the instrument meets the suggested or specified
regulatory performance.  It is another question, beyond the scope of this
handbook, whether "necessary" relates to the application to which the data will
be used.

    There are two kinds of performance specifications, those which can be
verified by simple Inspection testing and those which require unusual test
equipment and experience.  The former should be tested and the latter certified
by the manufacturer.   The manufacturer should have either performed these tests
on one or more samples of the model design or arranged for such tests to have
been conducted by some calibration facility.  In either case, a test report
should be available to any who require the documentation.  The cost of the copy
of the test report should not be much larger than the cost of normal copying.
If the manufacturer does not provide such documentation, the choice is between
accepting the manufacturer's unsubstantiated claim or having a specific test
run.

    A good QA Plan will provide for a QA sign-off of the procurement document
in order to assure that equipment capable of the required performance is being
purchased and that the capability can be verified.  Purchasing by brand name is
often expedient where the performance of a model has been verified and all that
is required is more units.  This practice is also cost effective when
considering spare parts and instrument technician training costs.

    The parts of wind speed and wind direction sensors which predictably
deteriorate and seriously influence the performance of the sensors are the
bearings.  It is acknowledged that an experienced Inspector can "measure"
bearing condition by feeling or spinning the shaft.  The receiving inspection
is a protection against putting defective equipment Into the field.  It Is not
a necessary link in the documentation trail for data validity purposes.  True
torque measurements for data validity will be most valuable at th
-------
                                                             Section No. 4.2.3
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page;     2  of  5
             Table 4.2.3.1  - Anemometer Performance Specification
Range

Thresholdll2(t)

Accuracy (error)
                            1,2,
            Distance Constant (t)
             0.5 to 50 m/s

             O.S m\s

             (0.2 m/s + 5% of observed)
                            3
             5 m at 1.2 k'g/n  (standard
             sea level density)
               (t) as determined by wind tunnel tests conducted
                   on production samples in accordance with
                   ASTM D~22.ll test methods (ASTM, 1985a).
               ($} aerodynamic shape (cup or propeller) with
                   serial number to be accompanied by test report,
                   traceable to NBS, showing rate of rotation vs.
                   wind speed at 10 speeds with 0.1 m/s resolution.

                1  subject to receiving inspection
                2  transducer with signal conditioner
                4.2.3.1.1  Threshold

                The threshold receiving test should examine the system output
with the anemometer not turning (below sensor threshold) and with the
anemometer turning at an equivalent 0.5 m/s.  If the cup is a Cllmatronics F460
Vinyl Cup Set (100083) or Heavy Duty Cup Set (101287),  the published constant
for mph and 30 pulses per revolution is 9.511 which converts to 1.41 meters per
revolution.  At 0.5 m/s the cups should be turning at

                            0.5+1.41  =0.35 rps,
or one revolution in 2.8 s.  The cup assembly can be turned by hand to
approximate that rate of rotation.  If the anemometer is a Young propeller
(08234, 18 cm polypropylene), the turning factor is 0.294 meters per
revolution.  At 0.5 m/s the propeller should be turning at
or two revolutions in 1.2 s.
propeller by hand.
0.5 + 0.294 * 1.70 rps,

  This also can be approximated by turning the
                The key measurement for threshold, however, is starting torque.
This requires knowledge of the K value (see 2.1.1.2) which should be available
from the manufacturer.

                4.2.3.1.2  Accuracy

                The receiving test for accuracy is the conversion of rate of
rotation to output in units of wind speed.  The transfer function, supplied by
the manufacturer, should be In terms of rate of rotation (rps) vs. wind speed
(mps).  The receiving inspector simply turns the anemometer shaft at a few
known rates of rotation to see if the systemoutput compared to the predicted
output is within the tolerance specification.

-------
                                                             Section No. 4,2,3
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page;     3  of  5
                4.2.3,1.3 Distance constant
                The distance constant determination requires a special wind
tunnel test and is beyond normal receiving inspection capability.  The time
constant of the anemometer circuitry will influence the effective system
performance.  Assume the manufacturer's value for distance constant of the
sensor is one meter.  At a wind speed of 10 m/s the sensor will have a time
constant of 0.1 s.  Assume the time constant of the system electronics is 2 s.
Then, at a wind speed of 10 m/s, the system time constant Is 2 s or 20 m.  In
this example, the system electronics would need a time constant of 0.1 s or
less if the response capability of the sensor is to be fully available.  At a
wind speed of 0.5 m/s the sensor responds in 2 s, the same as the system
electronics.  In this example, the system electronics dominates the sensor
response at all speeds.   If the sensor response is to be available at all
speeds {up to 25 m/s), the electronics time constant must be 0.04 s or less.
The time constant can be measured at the receiving test by timing how long it
takes for the output to reach 63.2% (1-1/e) of a step change in speed.  The
step change can made by turning the anemometer shaft at a known rate of
rotation and then Instantaneously stopping its rotation.

       4.2.3.2  Wind Direction

       An example of a wind direction specification is found in Table 4.2.3.2
(see EPA, 1987b).  This performance specification is also hypothetical but it
is one which will meet the requirements of EPA (1987a, PSD).  Each instrument
includes a sensor, signal conditioner, and recorder.  When an attribute of the
sensor is affected by the signal conditioner or recorder, a keyed comment will
be made.
              Table 4.2.3.1 - Wind Vane Performance Specification
              Range


              Threshold1>2(t)
                              1 2
              Accuracy (error) ' (t)
              Delay Distance (t)
              Damping Ratio (t)

              Overshoot2(t)
                          001 to 360 degrees or
                          001 to 540 degrees
                        £0.5 m\s

                        s 3 degrees relative to the
                            sensor mount or index -
                        2 5 degrees relative to
                            TRUE NORTH

                        s S m at 1.2 kg/m  (standard
                            sea level density)

                        a 0.4 at 1.2 kg/m3
                        £ 25% at 1.2 kg/m
(t) as determined by wind tunnel tests conducted on
    production samples in accordance with
    ASTM D-22.11 test methods (ASTM, 1985a).
 1  subject to receiving inspection
 2  transducer with signal conditioner

-------
                                                             Section No. 4,2.3
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page;     4  of  5
                4.2.3.2.1  Threshold
                The threshold receiving test is a starting torque measurement
(see 4.2.2.2.1).  To relate the torque measured to wind speed and off-set
angle, a K value is required, either from the manufacturer or from an
independent test.  The torque measurement may be made with the vane assembly
removed or with the vane assembly in place.  If the latter is chosen,
verticality is essential to negate any out-of-balance in the vane assembly from
biasing the test.  Also there must be no air motion.  Very small air motions
will bias the test.  Use a smoke puff to be sure the air is still and refrain
from breathing in the direction of the vane surface.

                4.2.3.2.2  Accuracy

                The receiving inspection is the best place to establish the
true non-linearity, if any, of the direction vane transducer.  A test using
some circle dividing fixture capable of fine resolution, 1 deg. for example,
will provide a record which can be referenced in future field spot checks.
Without such a test it is hard to prove accuracy of £ 3 deg.  If several units
show the same pattern of non-linearity, it should be acceptable to sample
future units and accept a generic shape of the error.  When a long series of
samples is planned, there is a tendency to devise methods which are quick.  The
time constant of the signal conditioning circuit must be known to establish the
minimum time between position change and output reading or recording.  If the
step is small, like 1 deg., three time constants will deliver 95% of 1 deg.,
which is good enough.  If the step is large, like 180 deg., three time
constants will deliver 95% of 180 or 171 deg, which is not good enough.  It
takes seven time constants to deliver 99.9% or 179.8 deg.

                The receiving inspection cannot include the orientation error.
The manufacturer does not deliver orientation.  There may be orientation
fixtures, however, which assume that an optical centerline is parallel to the
line set by an orientation pin.  This assumption can be tested.  Field
orientation may be based on the orientation of a crossarm with the assumption
that the output angle when the vane is parallel to the crossarm is known.  This
assumption can be tested or the alignment fixture set in laboratory conditions
to the desired output.

                4.2.3.2.3  Delay distance and overshoot

                These dynamic characteristics require a special wind tunnel
test and their determination  is beyond normal receiving inspection capability.
The time constant of the wind direction circuitry can influence the system
performance as it can with wind speed.  Assume the manufacturer's value for
delay distance is one meter and the time constant of the electronics is 2 s,
At wind speeds of 10 m/s the  time constant of the wind vane is 0.1 s and at 0.5
m/s it is 2 s, the same as the system electronics.  For this example,
therefore, at all speeds above 0.5 m/s the performance of the wind vane is
being limited by the time constant of the electronics. The time constant can be
measured at the receiving test by timing how long it takes for the output to
reach 63.2% (1-1/e) of a step change in direction.  For example, the step
change can made by quickly turning the vane from 000 to 180 deg.  The  time
constant is the  time required for the system output to change from 000 deg. to
113.8 deg.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    "
                                                             Date:   17 Sep S?
                                                             Page:    5  of  5
       4,2.3.3  Measurement Sys tem
       All the elements of a system of signal conditioners, recorders and
monitors will require checking for correct function.  The receiving inspection
should include testing these various sub-systems.  There may be a calibration
switch which replaces the sensors with simulated conditions.  Assume a systeo
has a calibration switch which substitutes the equivalent of 25 ra/s and  180
deg. to the input of the signal conditioning boards.  In 4.2.3.1 and 4.2.3.2
the sensors were providing the input values.  It Is possible for these tests to
show perfect results and yet the outputs with the calibration switches on could
show the system to be out of calibration.  The reason would be that the
adjustments for the substitute calibration inputs were off.

       After the calibration inputs have been adjusted and the "output"  shows
the system to be in calibration, a parallel analog recorder may show incorrect
values.  This event could be caused by an Incorrect adjustment in the interface
which drives the analog recorders from the output.  So far the "output"  is
assumed to mean the voltage which goes Into the data logger and becomes  the
archived data.   There may be monitoring meters or digital displays on the
system panel.  These monitoring meters may differ from the "output" because
they have Individual adjustments.  All the sub-systems should tell the same
story and the receiving inspection should verify that they do.  In fact, it is
rare when a system arrives In receiving with the various outputs in
disagreement, but they must be checked.

       It goes without saying that the receiving function records the model
numbers and serial numbers of the component parts and checks the parts received
against the purchase document and the shipping document.

-------
                                                             Section No. 4.2.4
                                                             Revision. No.    j
                                                             Date:    17 Sep S9
                                                             Page:     1  of  5


4.2.4  INSTALLATION

       4.2.4.1  General Considerations

       From a QA point of view, there are aspects of the installation which
should be considered.  Perhaps the most important of these is siting.  See
4.0.4.4 for general siting criteria and discussion.  From a QA point of view,
however,  failure to meet the recommended siting criteria may be necessary.  If
the general site is selected for other measurements for good reason, the wind
sensor siting may be only a best compromise.  There are considerations which
set the options for the compromise.  Among these are technical and budgetary
considerations.  The qualitative judgments which go into siting are briefly
discussed here.

       If buildings or trees are likely to interfere with the wind speed or
direction sensing,  try to locate the tower or pole such that the wind sensors
will most faithfully record speed and direction for the direction of primary
concern,  e.g. for directions that would take an effluent toward a residential
area.
       Another important technical consideration is accessibility of the
sensors.   There should be no hesitation in taking a hands-on look at the
sensors whenever a performance question arises.  Yet this is the most difficult
task at most sites.  Some sites require special "climbers" to retrieve a sensor
and wait to return it to its installed position.  These field people jmay not be
trained to handle delicate instruments.  It may be so difficult or expensive to
get a sensor down that suspected bad data will be accepted rather than facing
the problem.  This reluctance causes mounting hardware to become corroded to
the point that the sensor cannot be removed without damage.  Most of the
breakage of delicate sensor parts results from handling while climbing on a
tower.  If the direction sensor does not have an alignment fixture, it may not
be possible to remove the sensor without going through the orientation
procedure at re-installation.

There are several ways to overcome most of these problems.  They all require
design forethought in installation.  First, the sensors need to be easily
removed from the mounting structure.  They need to be as easily connected to
the rest of the measurement system when they are down, either with th* same
cable or a suitable substitute cable.  One popular cup and vane design uses a
crossarm which contains non-removable cabling.  Either the whole crossarm
assembly must be taken down with the sensors or there has to be a substitute
crossarm to plug the sensors into at ground level.  Some towers, the siiort 10 •
types, can be tilted to access the sensors while still connected.  In this case
the sensors are about 90 deg. from their operating position.  Some tests
require the sensors to be vertical.  In these cases the sensor still needs to
be removed and re-connected.  Ideally, the crossarm is left exactly as is so
when the sensor is replaced, nothing physical has changed with respect to
verticallty or orientation.  Some towers telescope for access to the top where
wind sensors usually are mounted.  This makes access easier but removal and
reconnection is still necessary.  Some towers have elevators which transport
the sensors up and down the tower.  How they deal with cables depends on the
elevator design.

       An ideal installation is one where the operator can get hands on the
sensor, perform a test adjacent to the electronics and recorder, and re-install

-------
                                                              Section  No,  4.2.4
                                                              Revision No.     3
                                                              Date:    17 Sep 89
                                                              Page:     2  of  S

 the  sensor,  safely,  alone and within one hour.   There  are  no  technological
 reasons  why  this  cannot be done,  except for tall towers where an  elevator may
 take 20  minutes for  a one-way trip.   There probably is not a  budgetary reason
 for  avoiding something like this  ideal installation, once  the cost of invalid
 data and true operating costs are factored into the formula.   It  is usually not
 done simply  because  the need for  service is overlooked and the method of access
 is not pre-planned.

       4.2.4.2 Wind Speed

       The wind speed sensor is most susceptible to error  from shadowing and
 interference.  Aside from the need to have the anemometer  properly exposed, the
 only other consideration is verticality (for cup anemometers).   If the cup
 wh^e]  is well balanced, a small angle (1 deg.  or less) in  mounting is not
 important.   If the cup wheel is not well balanced,  the starting threshold will
 be degraded.

       4,2,4.3 Wind Direction

                4.2.4.3.1  Exposure

                The  problem with  verticality for the direction vane is just the
"same as  with wind speed.   For a well balanced vane assembly,  a small  angle  fro«
 vertical is  not important.  If the vane assembly is not well  balanced, the
 starting threshold is raised and  a predominant direction for  light winds may
 not  have any basis in fact.

                Wind vanes are often used for dispersion applications by
 calculating  the standard deviation of the wind direction about the mean
 direction,  sigma  theta.  Unless the wind vane is at the tower top, there will
 be some  direction where the wind  goes through and around the  tower before it
 gets to  the  vane.   The farther from the tower the wind vane is mounted the
 smaller  the  sector with tower interference.  The interference sector  can be
 selected by  placing the vane on the appropriate side of the tower.

                4.2.4.3.2  Orientation

                Of all the sources of error for a wind direction measurement,
 the  orientation of the vane to TRUE NORTH has the potential and reputation of
 being  the largest.  A bad orientation provides a fixed bias to the data which
 can  be removed.   If the vane is moved and the constancy of the bad orientation
 is in  question, the data may not  be recoverable.  The  method  of wind  vane
 orientation  must  be capable of 1  deg. accuracy with 2  or 3 degrees as the upper
 limit  of the error.   Two steps are necessary to achieve an oriented wind vane,
 First,  the location of TRUE NORTH must be found to an  accuracy of less than 1
 deg.  Secondly, the wind vane "index" must be aimed at that location  with an
 accuracy of  better than 2 deg. (see 4.2.2.2.2.3 for a  discussion of the
 location of  the normalized error  "index").

                TRUE NORTH as distinguished from magnetic  north is usually
 found  by reading  a magnetic compass and applying the correction for magnetic
 declination.  The declination can be read from a USGS  map.  The Fox Island
 Station declination, according to the 1959 (revised in 1981)  map, is  20.5°.
 The  USGS Is  now providing a computer service called GEOMAG.  See below;

-------
                                                             Section No. 4.2.4
                                                             Revision No.    ;
                                                             Date:   17 Sep 8=
                                                             Page:    3  of  5
CONNECT
        Unauthorized use of this U.S. Government computer system
                     is punishable under PL98-473
Welcome to the USGS Branch of Global Seismology and Geomagnetism
               On-line Information System
Type Q for Quick Epicenter Determinations (QED3
     H for Historical Epicenter File Searches  (EIS)
     M for Geomagnetic Field Values
Enter program option (Q, H or M): m
                                  GEOMAG
The International Geomagnetic Reference Field  (IGRF) was revised in August,
1987.   The models for 1945.0, 1950.0, 1955.0, and 1960.0 have been superseded
by new definitive models (see, for example,  EOS Transactions, American
Geophysical Union, vol. 69, no. 17, April 26,  1988, pages 557-558). The new
models were installed on June 21, 1988.  Please note that the revision affects
field values for dates between 1945.0 and 1965.0, but not those for later
dates.
Problems or suggestions?  Please contact Norman Peddle, U.S. Geological Survey,
MS-968, Federal Center, Box 25046, Denver, CO 80225. Telephone:  (303) 236-1364
(FTS 776-1364).
Press RETURN to continue:
Do you want information about this program (Y/N)?      [ ]  N
Options:   1) Field Values (D, I, H, X; Y, Z, F3
           2) Magnetic Pole Positions
           3) Dipole Axis and Magnitude
           4) Magnetic Center                          [1]   [ ]
Display values twice (Y/N)?                            [Y]   [ ]
   Name of field model:                    [         ]          ?
The following field models are available:
     Name
   IGRF85
   USCON85
   USALA85
   USHAW85
Name of field model:
   Date:
Latitude:
North or South  (N/S):
Longitude:
East or West (E/W):
Elevation:
       Type
Spherical Harmonic
Spherical Harmonic
Spherical Harmonic
Spherical Harmonic
                     i
  Date range
1945.0 - 1990.0
1985.0 - 1990.0
1985.0 - 1990.0
1985.0 - 1990.0
       1
     Region
   World
   48-States
   Alaska
   Hawaii
USCON85
                     [1/25/891

                     [ 1  N
             J47{.25

             1122.6292
                     [ I  W
                     10.0]
    [250 feet
Model: USCON85
Date  : 1/25/89
   Latitude : 47.25    N
   Longitude: 122.6292 W
        Elevation:   250.000 ft

deg
19
19
Annual
0
0
D
rain
47.9
47.9
change :
-5.5
-5.5

deg
69
69

0
0
I
mln
29.0
29.0

-0.7
-0.7
H
nT
19526
19526

-1.9
-1.9
X
nT
18372
18372

8.7

Y
nT
6613
6613

-29.8

2
nT
52181
52181

-39.6

F
nT
55715
55715

-37.7


-------
                                                             Section No. 4,2.4
                                                             Revision No.    ~
                                                             Date:   17 Sep £9
                                                             Page;    4  of  =

                The GEOMAG program was accessed by calling 1-800-358-2663
through the modem of an "XT clone" using a "ONE TO ONE" communications program.
Using the capture feature of ONE TO ONE, the following communication was
recorded.   Note that the Fox Island Station declination for 1/25/89 is 19.8° cr
0.7° less than the map indicated.  This is consistent with the 0 deg. -5.5 min.
annual change for the roughly eight years since the map was revised.

                The other way to find the direction to TRUE NORTH employs some
astronomical observation.  While the compass method is clearly easiest,  it is
also the most prone to error.  Good training and equipment will reduce these
errors to an acceptable level, but not the "less then 1 deg." advised above.
Training will minimize errors from the influence of nearby metal objects and
the mis-application of the declination correction, but local variation in the
isogonic field is unknown.  On the other hand, the observation of astronomic
bodies can be unambiguous.  Polaris, the north star, will provide TRUE NORTH to
within 1 deg. (without correction) on any clear night.  The true solar noon
method will provide the north-south direction to within 0.1 degree on any clear
day, given the station longitude, date and an accurate clock.  A simple Basic
program will provide the azimuth angle to the sun at any time of day given the
station longitude, latitude and date.  Examples of the two solar methods are
given below,

                     4:2.4.3.2.1 True Solar Noon Method

                     The True Solar Noon (TSN) method finds the time at  some
particular date at some particular longitude when the sun is in the north-south
plane passing through the North Pole, the South Pole and the longitude
selected.   If the sun is not directly overhead (elevation 90 deg. ) the azimuth
line to the sun is TRUE SOUTH or TRUE NORTH.  Two calculations are required.
First, find the time of the Local Apparent Noon (LAN) from the longitude.  The
examples shown here are for:
            Fox Island, WA  (Long. 122.6292, Lat. 47.2500), 07/04/90  and
            New Or leans, LA  (Long.  90.1100, Lat, 30.0000) 12/25/90.
T    =  12:00:00 +  4(Long.  -  15n),  where  n is  the number of time zones from
 LAN
Greenwich.  Table 4.2.4.1 is a list of n values for United States time zones.

                           Table 4.2.4.1  Time Zones
Time Zone
Eastern
Central
Mountain
Pacific
Yukon/ A 1 aska
Hawaii
n
5
6
7
a
9
10
T   (WA) = 12:00:00  + 4(122.6292 - [15 X 8j) = 12:10.52 = 12:10:31  PST
 LAH
T   (LA)  = 12:00:00 + 4(090.1100 -  [15 x 6]) » 12:00.44 * 12:00:26 CST
 i» AN

Secondly, correct for the  Ephemeris of  the sun.
T    =* T    - A, where A Is the correction found in Table 4,2.4.2.
 TS n    LAN

-------
                                                             Section No, 4.2.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  S
                      Table 4.2.4.2  Ephemeris of the Sun
From the Nautical AJmanac - 1989 Yachtsman's Edition
Equation
Date of time
m. s.
Jan. 1 - 3 24
4 -448
7 -608
10 -724
13 -835
16 -940
19 -1039
22 -11 31
25 -12 16
28 -12 54
31 -13 25
Feb. 3 -1349
6 -1405
9 -1414
12 -1416
15 -1411
18 -1400
21 -1342
24 -13 18
27 -1249
Mar. 2 -12 15
5 -11 37
8 -1055
11 -1010
14 -921
17 -831
20 - 7 38
23 -644
26 - 5 49
29 -454

Equation
Date of time
m. s.
Apr. 1 - 4 00
4 -307
7 -216
10 -126
13 - 39
16 + 6
19 +48
22 +1 26
25 + 1 Si
28 + 2 29
May 1 -i- 2 53
4 +313
7 +328
10 4-338
13 +342
16 +342
19 +337
22 +326
25 +311
28 + 2 52
31 + 2 28
Jun. 3 + 2 00
6 +129
9 +55
12 + 19
15 - 19
18 - 58
21 - 1 36
24 -215
27 -253
30 - 3 30
Equation
Date of time
m. s.
Jui. 3 - 4 05
6 -437
9 -506
12 -532
15 -553
18 -609
21 - 6 20
24 - 6 27
27 • 6 28
30 -625
Aug. 2 -615
5 -601
8 -540
11 -515
14 -444
17 -408
20 - 3 28
23 -243
26 - 1 55
29 - 1 03
Sep. 1 - 8
4 +50
7 +1 50
10 +252
13 +355
16 +500
19 +604
22 +708
25 +811
28 +912

Equation
Date of time
m. s.
Oct. 1 +10 11
4 +11 08
7 +1202
10 +1252
13 +1338
16 +1420
19 +1456
22 +15 27
25 +15 52
28 +1610
31 +1621
Nov. 3 +1625
6 +1622
9 +1612
12 +1554
15 +1528
18 +14 55
21 +14 14
24 +13 26
27 +1231
30 +11 29
Dec, 3 +10 21
6 +908
9 +751
12 + 6 29
15 •*• 5 04
18 +-337
21 •+ 2 08
24 -1-38
27 - 51
30 - 2 20
   T   (WA)  = 12:10:31 - (-4:16)  « 12:14:47 PST
    X5N
               13:14:47  PDT
   T   (LA) « 12:00:26 - (-0:08)
    TSN
12:00:34 CST
                Once  the time of TSN is known, all that reaains is to observe
the position of the line to the sun at TSN.  An easy way is to use a loosely
mounted theodolite set at 180 deg. to track the sun.   [CAUTION: EYE DAMAGE MAY
RESULT FROM LOOKING AT THE SUN WITHOUT SUITABLE PROTECTIOM - Remember that
harmful UV rays can be present when visual light "looks" safe - USE AN EYE
SAFE FILTER]  When a  watch (one second resolution set to WWV or equivalent
source for the correct time) shows TSN, tighten the theodolite mounting.  At

-------
                                                             Section No. 4.2.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  8

that instant the sun is in the cross hair of the theodolite and the theodolite
correctly labels the azimuth angle as ISO.  Once set, the theodolite can be
used to find the bearing to any distant feature which might be selected as an
orientation target.  Another method Is to mark the end of the shadow of a
vertical tower at TSN, thus establishing a N-S line from the base of the tower
to the mark.

                     The two drawbacks to the TSN method are weather and
schedule.  If the sun is obscured at TSN the observation cannot be made.  Also,
if other activities command higher priorities, the time of TSN might not be
available for the sighting.

                     4.2.4.3.2.2 Solar Azimuth Method

                     The azimuth angle to the sun can be found at any time if
the latitude is also known.  A Basic program (Blackadar, 1985) which contains
the necessary subroutines has been edited to provide the outputs shown  in
Figure 4.2.4.1.  These are the same two examples as are used in 4.2.4.3.2.1.
The program listing is given as Figure 4.2.4.2.  Notice in Table 4.2.4.3 that
the two methods do not agree.  The differences are trivial.  Even at the fast
angular motion of July, the sun moves about 0.5 deg. per minute.  The roughly
quarter minute difference In methods represents only a little over 0.1 deg.
uncertainty.  Notice also the nonlinearity difference between winter and summer
which makes simple extrapolation impossible.

                     A Brunton compass, mounted on a tripod, can be used for
solar sighting.  The mirror can be set to project the sun and the sighting
points and lines on a white piece of paper.  The compass needle can be used as
a reading index or an additional protractor and pointer can be added to the
compass mounting hardware.

-------
                                                                     Section  No. 4.2.4
                                                                     Revision No.     0
                                                                     Date:    17 Sep 89
                                                                     Page:     7  of  8
MY? 4
MONTH? 7
YEftR? 1990
Ttie
H8,H1N? 11,30

HR,«H? 12,00

HR,HI»? 12,30

HR,««? 13,00

HfUKIN? 13,30

HR.HIN? 13,00

Elevitic

57.81

61.34

64.00

65.43

65.42

65.42
                      SOLAR ftZlHUTH ANGLE
                      lEDNESDftT 4 JUL 1990
                Fox Island, 1A
                Loniitude 122,4252   Latitude 47.25
                lay of Year 185     Julian Diy 244807?
                  Transits Meridian 13  14 53 PDT
                                 130.15

                                 141.90

                                 155.86

                                 171.73

                                 188.40

                                 171.73
   BAY? 25
   KKIH? 12
   YOU? 1990
                                                            use
                          AZOlflH
                      TUESDAY   25 Da 1990
                fe* Orleans, Lft
                Loiaitu* 90.11     Latitude 30
                Say of Year 359     Julian Day 244=3
                  Transits Iferidian 12 0 34 CST

                      Elwition  Aiiiuth

Ht,KH? 11.00

m.nm 11.30

Ht»H!h" 12,00

HR,HIH? 12,30

HR.BK? 13,00

2.35

S4.fi?

36.11

36,61

34.14

34.73
153.28

163.07

171 TT
i.i *»**<*'

179.86

188.38

196.65
Figure 4.2.4.1   Screen printouts for two azimuth examples
                   Table 4.2.4.3  Solar Method Comparisons
         Lo.La.D          WA  (7/4/903

         Method     TSN      Almanac  dif.

         Units      PDT       PDT     (s)

         TSN     13:14:47  13:14:53  -6
         LA (12/25/90)

   TSN     Almanac   dif.

   CST        CST     (s)

12:00:34  12:00:34    0
              where  Lo is the  station longitude
                      La is the  station latitude
                      D   is the  date of interest

-------
   ,50 STAS.-MSi; Fox W«^;WA^R£M STATION NAME  __ f
   '60READLO.IA .................. * .......................        """""
   100. DATA122.629,4735:REMLQNG;rrUOE.& LATITUDE
   1 10' READ Pt.OBXb,Ll>AAt£C^6 ..............
   120 DATA 3.l4lS926S4,.40B095.4A83766ia,.01 7202781
   130 DATA 6.23471229.,017201670>01fi728..QQ218
   1 40 TR.PI/180-.FC.2Tl:REM TO RADIANS; FULL CIHClf
   iso suis'iNTft-O/is+j^REM STANDARD LONGITUDE
   180 TZ«Sl/1S-4: REM SELECTS TIME ZONE LABEL
   190 LO-LOTfttJUUrmSUSLTR
   210D1S.*SUNDAY  MONDAY  TUESDAY WEDNESDAY"
   220 D2$-THURSDAY FRIDAY  SATURDAY"
   230 DJ-01 J+D2S : X$m' - •
   240 MS-'JANFEBMARAPRMAYJUNJULAUGSEPOCTNOVDECT
   250 Z$-"ASTESTCSTMSTPSTYSTASTADTEDTCDTMDTPDTYDTADP
   260 TN-LO/FC+.S:REM LONGfTUDE TIME OFFSET + 12 HR
   270 PRINT "DAY-niNPUT D
   280 PRINT "MONTH";:INPUT M
   2SO IF M»12 THEN PRINT 'INVALID DATE'rGOTO 280
   300 PRINT •YEAH'slNPUT YR
   310 X.1:Y-1:GOSUB 2410
   32B JI-TflEM YEAR DAY 1
   .130 X-O:Y-M.-GOSUB 241 0
   340 YD.T-J1 +1 AEM DAY OF YEAR
   360 WD-lNT(r(X-YH.I):REM DAY OF WEEK
   380 T-T-K344B.5+TN-.HEM T IS NOW TIME OF LOCAL MEAN NOON
   385 DT-.00059+2iE-08T : T-T+OT : REM EPHEMEHIS TIME
   3SO PRINT TAB(28)rSOLAR AZIMUTH ANGLE"
   405 PRtNTTAa(28);
   410 PRINT MlO$PJ,fWO+1.fl):
   420 PRINT D#IID$(M$,3'(M*1 M.3KYR
   423 PRINTTAB{20);STAi;
   490 X-YD-WD:
   £00 IF X>60 AND X<298 THEN TZ.T2«-7 : SUY
   610 GOSUS 2860: REM FIND SUN AT LOCAL MEAN NOON
   620 IF OE>PITHEN DE-OE-FC
   S30 Q-ML-RAflEM EQUATION OF TIME {NOT DISPLAYED)
   640 03-OE : REM SAVE DECL FOR HEAT BUDGET
   680 X— .0145439 : GOSUB 2360
   690 tF ABS(Y)«1 THEN 720
   710 GOTO 780
   720 SO-T-fl +L1/FC) : H-SO : GOSUS 2260
   72S TC-.00274'SCrSIN(O8)'COS(TD"SW(LA)
   730 Z»SIN(SO)'OOS(LA!*{COSPI THEN H.H-FC
                                        2230 RETURN
                                        2250 FOR U.I TO 5
                                                   ) : Y.COS(ZT) : GOSUB 80tO
                                        Z27S MLAO*LVfr.TN+{Sl.»ZyFC) ; NEXT U
                                        2280 ZT-Z: RETURN
                                        2310 F X<0 THEN X-X**C : GOTO 2310
                                        2315 W«X*24/FC : X-1NT{W)
                                        2320 Z^W-X)*60 : Y-(NT£Z)
                                        2330 ZJHT((Z->TSO) : RETURN
                                        2370 F ASSfl>1 THEN 2390
                                        2380 X-SQR(1-Y*2) : GOSUB 2010
                                        2390 RETURN
                                        2410T-3ST(YR.1880)
                                        2420 T.T-WTCT fm+lNT({Y*i)n23)'*)
                                        2430 S-SGN(Y-9)*-ABS+1-25>EC*2-SIN(2'MA)
                                        2910 TAOW*OL:TL-ML+OLREM TRUE ANOMALY & LONGITUDE
                                        2920RV-{1.eC*2yCI+E(rCOSfrA5)SEM RADIUS VECTOR
                                        2930X-SK{TLrS!N
-------
                                                             Section No.  4.2.5
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  4
4.2.5  CALIBRATION
    Calibration, as defined on page 3 of the Purpose statement in the beginning
of this handbook,  qualifies the process as both a measurement and adjustment,
if necessary, of the performance of the system and Its components.
Manufacturers usually include in their manuals the details of all the available
calibration or adjustment points.  From a QA standpoint,  the important
consideration is how the system is working as a whole.  Since only parts of the
system are adjustable, the relationship of these adjustments to the whole
system must be known.  This brief section will focus on documentation of
calibrations and methods to verify the system response to subcomponent
adjustments,

       4.2.5.1  Wind Speed

                4.2.5.1.1  System accuracy

                The part of a calibration which challenges the entire system,
except for the coupling or reaction of the sensor to the wind,relates the rate
of rotation of the anemometer shaft to output speed.  It does not matter if the
rate of rotation is caused by a synchronous motor or a d.c. motor with a
provision for shaft revolution counting.  What does matter is the accuracy of
the determination of AVERAGE rate of rotation and a common averaging PERIOD
used by the system and the challenge.  The operators may choose to conduct this
calibration with the sensor installed on the tower.  When multiple outputs
exist, the calibration should record values from each of them, but the critical
output is the one used to produce the official archieved data.


                The accuracy determination depends on both the method used in
the challenge and the accuracy of the measurement of the input.  If a
synchronous motor is used, there must be some reason to believe the motor was
turning in sync with the commercial power.  Repeated samples which do not
change is one form of evidence.  Commercial power is generated within a
frequency tolerance of 60 ± 0.1 cps.  Synchronous motors which are hand held
with a flexible coupling to the anemometer shaft may go in and out of sync
providing a slightly changing output.  Shaft rotation counters can also produce
erroneous outputs.  Some evidence of their performance, such as counting a
synchronous motor shaft rotation or simply counting revolutions at a slow rate,
is needed in the documentation of the test equipment, preferably before and
after field use,
                4.2.5.1.2 Component accuracy

                If the system has built in calibration circuits, they should be
calibrated at the same time as the total system.  They are handy to use on a
routine service schedule, but there needs to be some evidence of their
calibration.  If panel meters or portable DVOMs are used to check the signal
conditioner or transducer sub-system, there needs to be evidence that they are
in calibration.  It  is possible to adjust a circuit to provide the required
output on a meter which has a 2% error and thereby introduce a 2% error to the
calibrated system output.

                The  calibration of the sensor starting threshold can only be a
measurement.  Adjustment is usually impossible.  The exception might be the
amount of end play in the shaft-bearing assembly, but that level of sensor

-------
                                                             Section No.  4,2-5
                                                             Revision No.    "•
                                                             Date:   17 Sep 85
                                                             Page:    2   of  4

repair is usually  left  to  the  laboratory or shop for good reason.  The accuracy
of the torque measurement, or  non-measurement, is also important.  Assume a
torque watch, or similar device, with a range of 0.003 to 0.030 oz-in.  The
threshold of measurement is 0.003 oz-in or 0.22 g-cm.  If a cup anemometer has
a K value of 1.4 (see 2.1,1.2), the torque provided by a 0.4 m/s  (0.9 mph) wicd
is 0.22 g-cm [from T=Ku2].  The torque provided by a 0.5 m/s (1.1 mph} wind is
0.35 g-cm.  If the torque  watch cannot measure the starting torque of the shaft
because it turns before the indicator moves, the starting torque  is < 0.22 g-cm
and the starting speed  is  < 0.4 m/s.  If, instead, the starting torque reads
0,35 g-cm (0.005 oz-in), the starting speed is 0.5 m/s and within
specification.  If the  starting torque reads 1.0 g-cm (0.014 oz-in or about
half scale on the  torque watch), the starting speed is 1.4 m/s and clearly in
need of service.   Documentation of this measurement will tell the data QC
inspector that the data from this anemometer is in error in the indicated 0.2
(assuming a 0-2 m/s offset) to about 3 m/s range.  (3 m/s wind provides 12.6
g-cm torque or about an order  of magnitude more than that provided at 1 m/s.
The difference between  0.35 and 1.4 is not likely to be visible at 12.6)  The
true wind speed will oe higher tnan the indicated speed.  At indicated 0.3 m/s
it would be 1.5 m/s and at indicated 3 m/s it would really be 3 m/s.

       4.2.5.2  Wind Direction

                4.2.5.2.1  System accuracy

                The system calibration of a wind vane can be checked on the
tower by aiming the vane to and from known directions, such as a distant
mountain peak or similar feature.  If checks are made with respect to a mounted
component, such as a crossarm, the orientation of the crossarm also needs to be
checked.  A single distant feature should be the orientation target with  a
known bearing with respect to  TRUE NORTH.  Other targets can be secondary
checks which challenge  both the orientation and the performance of the
transducer.  For systems using the 540 format, the targets should be reached
after a clockwise  revolution and then again after a counter clockwise
revolution to challenge both parts of the transducer.

                Before  the transducer is removed from the tower, a
documentation of the as-found  output with the vane held pointing at or from the
orientation target is essential.  This single act provides the basis for  data
validity for the period beginning with the previous as-left record and ending
with this as-found reading.  Since the sensor should not have been removed fro«
the tower or adjusted in orientation without the as-found and as-left readings,
these values should be  the same, within the capability of pointing the vane (1
deg., 2 deg. at the most).  If they are not, the data QC inspector will have
some detective work to  do.  Usually, when the sensor is removed and used  in a
calibration at the location of the rest of the system, replacement in a keyed
fixture will cause the  as-left value to be the same as the as-found.  Figure
4.2.5.1 shows three examples of how manufacturers provide an orientation  key
for wind direction sensors.  If there is no keyed fixture, the full orientation
procedure will be  required.

-------
                                                             Section-No. 4.2.5
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  4
Figure 4.2.5.1  Methods for keying orientation of direction sensors

                One simple orientation procedure requires a clamp which will
hold the vane from turning.  A hose clamp will work for some designs.   Figure
4.2.5.2 shows a hose clamp used for this
purpose.  Tape which does not stretch is
marginally useful.  Stretchy tape like
duct tape or electricians tape will only
work on a perfectly calm day.  Set the
vane so that the output is the correct
value for the orientation target (see
4.2.2.2.2.1).  If the angle of the
orientation target is coincident with a +2
deg. error relative to the average error
of 0 deg., the output should be 2 deg,
higher than the bearing of the orientation
target.  Only in this way will the relative
error of the sensor be distributed equally
about TRUE directions.  Tighten the clamp so the output is both correct and
constant.  Mount the clamped sensor on the tower and turn it until the vane
points at the orientation target.  Clamp the vane in place.   Verify that the
output is still correct before removing the vane clamp.

                4.2.5.2.2 Component accuracy

                The same comments regarding calibration circuits, parallel
recorders and panel meters apply to wind direction as they do to wind speed,
mentioned in 4.2.5.1.2 above.  With the sensor next to the signal conditioner
(attached with either the operating cable or a suitable substitute) and with a
fixture which holds known relative directions, the signal conditioner can be
adjusted if required.  The 540 offset voltage, if one is used, can be tested
and adjusted.  The output voltage vs. position can b'e set.  The open space In
the potentiometer, if one is used,  can be measured and adjusted for.
    Figure  4.2.5.2
A direction vane cl&mp

-------
                                                             Section No. 4.2.5
                                                             Revision No.    I
                                                             Date:   17 Sep 89
                                                             Page:    4  of  4

                A single potentiometer has an electrical range of something
like 355 deg. with a mechanical range of 360 deg.  If the transfer function of
relative direction to voltage output is

                                  9 - 360 x V

where 6 is the angle in deg. and V is the output in volts (0-1 V scale), and
the maximum "full scale" output, set by shorting the potentiometer wiper to the
high side of the potentiometer, is 1.000 V, a small error will have been set
into the system.  The error will be +1.4 % of reading.  At 355 deg. the output
will be 360.  At 180 deg. the output will be 182.5 deg.  This adjustment error
added to the linearity error of the potentiometer may be more than is
acceptable.  If Instead, the signal conditioner is set to output 0.986 V when
the vane is set to 355 deg., the output will be 355 deg. (360 x 0.986).  At 180
deg. the output will be 180  (assuming no linearity error).  All of the error
between 355 deg. and 360 deg. is in that 5 deg. sector.

                Is this acceptable for PSD (EPA, 1987a) applications?  The
"wind direction system error" which cannot exceed 5 deg. Is the error of the
averaged wind direction samples.  If the mean direction were 355 deg, with a
range of ±5 deg., and if the distribution were bi-modal with half the values at
350 deg. and half at 360 deg.» and if the output voltage remained at 0.986 V
between 355 deg. and 360 deg,, the average output would be 352.5 deg. , a -2.5
deg. error.  If the dead space were at 0 V, the output would cause the bi-modal
distribution to look like half 350 deg. and half 360 deg. producing the correct
average of 355 deg.  This is a maximum error estimate.  True distributions
would cause smaller average errors.  Even a wind averaging 357.5 deg. with a
range of ±2.5 deg (the vane  is always in the dead space), the error is 2.5 deg.

                The starting threshold of the wind vane is important to
accurate low wind speed directions.  The design of the vane along with the
off-set angle (or error tolerance) provides a K value.  The K value along with
the starting torque of the vane assembly provides a threshold wind speed.
Assume a 5 deg. error tolerance and a K value of 15.  At 0.5 m/s the available
torque is 3.75 g-cm.  At 10  cm out from the axis or rotation, a force of 1/3 of
a gram should move the vane assembly.  This is another threshold of the torque
gauge situation.  At 1 m/s the torque available is 15 g-cm and at 10 cm the
force is a reasonable 1.5 g.  At 1 m/s and 10 deg. error tolerance, K becomes
37.5, the torque available becomes 37,5 g-cm, and the force at 10 cm is an
easily measured 3.75 g.

-------
                                                             Section No. 4.2.6
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2


4.2.6  OPERATIONS, MAINTENANCE AND QUALITY CONTROL

       4.2.6.1  Operations

       The important aspects of operations, from the standpoint of quality
assurance, are planning (see QA Plan, Section No, 4.1.) and documentation
(Section No.  4.9,1),  The purpose of operations is to acquire valid data.  For
wind measurements, this requires frequent (weekly, if possible) visual
examination of the sensors.  This is not a "hands-on" examination but simply a
look at the active shapes, cups, propellers and vanes, to be sure there has
been no physical damage.  Sensitive wind instruments can be damaged by hail and
by birds.  The nature of an analog recording, if one is used routinely,  will
tell how the sensor is performing.  Routine entries in the station log will
provide the evidence of attention to support validity claims.

       Calibrations are a part of operations.  A member of the operating
organization needs to become the "expert" on how the measurement system works
and what it needs to continue "in control" performance.  Regularly scheduled
calibrations build the expertise and the documentation showing measurement
accuracy.  The frequency of calibrations is a variable.  For a new
installation, a calibration during the installation is necessary.  A careful
look at the first week of operation will find early failures.  If all seems to
be going well, a calibration check after a month is prudent.  If no problems
surface, a full calibration at the end of the first quarter is advisable.  For
some site environments and some applications quarterly calibrations are
recommended.   Semi-annual calibration is the minimum frequency. If problems are
found they must be documented and corrected as quickly as possible.  The
requirement of 90% joint frequency of valid wind and stability data does not
permit much down time.  The frequency of calibrations or calibration checks
should be determined by the performance of the instrument system.  If problems
occur, the week-month-quarter frequency should begin again.  When it is
demonstrated that the system is once again "in control," routine calibration
frequency (semi-annual or quarterly) can resume.

       4.2.6.2  Maintenance

                4.2.6.2.1 Routine and preventive maintenance

                The only routine maintenance required for the wind system
should be applied during routine calibrations.  Sensors exposed to the elements
need cleaning and protective lubricants applied to their mounting hardware.
When a sensor needs to be removed for close inspection or calibration and it
cannot easily be removed because set screws or nuts are locked to their threads
by corrosion, a failure in routine maintenance is the reason.

                If the system has supply requirements, such as ink and paper
for analog recorders or tapes and printer paper for digital recorders, the
timely servicing of these requirements is a routine maintenance task.

                Preventive maintenance must at minimum follow the
manufacturer's recommendations.  Considerable damage can result by ignoring
this guidance.  Some people like to oil anything that moves.  Sensitive wind
sensors require specific care if the threshold is to be maintained.

-------
                                                             Section No,  4.2.6
                                                             Revision No.    C
                                                             Date:    17 Sep 89
                                                             Page:     2  of  2
                4.2.6.2.2  Corrective maintenance
                Parts for wind systems are not to be found anywhere except fron
the manufacturer.  This is true at least at the sub-component and component
level.   When a part falls or wears out, the new part usually must come from the
manufacturer.  This may take a week or two depending on the part and the
manufacturer.  It is prudent to have spare parts on hand to cover some
predictable failures.  A component plug-in philosophy Is the quickest way to
correct failures.  If a bearing or a potentiometer fails in a sensor, a new
calibrated sensor is simply plugged in while the failed one is repaired.  If a
circuit card fails, a new calibrated card is plugged in while the failed one is
repaired.

                The next level of spare part strategy is the sub-component
level.   Critical and difficult to buy parts are stocked and used to repair
sensors or circuit cards.  Conventional sensors will always need repair at so»e
point in time, bearings and direction potentiometers usually.  Circuit cards
are becoming so reliable that maintenance is hard to anticipate.

       4.2.6.3  Quality Control

       The quality control (QC) of a data monitoring program is a loop driven
by routine inspection of the data for validity.  The data QC person should be a
meteorologist who is familiar with how wind data should look and what kinds of
variety are provided by the atmosphere.  Such an inspector will spot problems
before they are obvious to an observer who may be an expert in another field
but is not a meteorologist.  If a technically qualified QC inspector is not
available, the best compromise that is available must be nade.  It is very
dangerous in terms of lost time if no routine data QC function is followed in
the QA Plan, or if there is no QA Plan.

       When z problem is found by the data QC inspector, a discrepancy report
is issued which brings the operators into the data QC loop.  Their inspection
and corrective action is reported back to the QC inspector closing the loop.
Because of this QC loop, the measurement system can be operated "in control"
and valid data produced.

-------
                                                             Section No. 4.2.7
                                                             Revision No.    1
                                                             Date:   17 Sep 89
                                                             Page:    1  of 2C
4.2.7  PERFORMANCE AUDIT METHODS

       4.2,7.1  General Considerations

       A performance audit Is the determination of the instrument system
accuracy made with an independently selected method and by a person who is
independent of the operating organization.  To make this determination for wind
measurements, knowledge of the input conditions imposed upon the sensors is
required.  Given knowledge of these input conditions, the transfer functions
and the system's data handling method, the output can be predicted.  The
difference between the predicted output and the system output is the error of
the system or its accuracy.

       The methodology starts with the ways of controlling and/or measuring the
input conditions.  When controlled inputs are used, as should always be the
case for starting thresholds, anemometer rate of rotation vs. output and
relative vane position vs. output, the accuracy of the output is easily
determined.  Of course, the accuracy of the anemometer transfer function is net
a part of this determination.  When the input conditions are not controlled,
as with the collocated transfer standard (CTS) method, the accuracy
determination has a larger uncertainty.  The CTS method does challenge the
anemometer transfer function.  The best performance audit uses both methods
where appropriate.

       A performance audit must follow some written procedures.  Since the
procedures must be relevant to the design of the instrument or system being
audited, only general principles will be described below with some specific
examples.  The data from the audit should essentially fill out an audit form.
It is important, however, for the auditor to be sufficiently experienced to be
able to deviate from the procedure or the form when the pursuit of truth leads
away from the expected.

       The starting point of an audit form is the documentation of the who,
what, where, when, and how the audit values were acquired.

                4.2.7.1.1  Who

                The performance audit report form should contain a space to
identify the auditor.  The audit report which summarizes the audit findings
should report the names and affiliations of the operators of the system.

                4.2.7.1.2  What

                The form should contain a section to identify the instrument
being audited by manufacturer, model number, and serial number.
Sub-assemblies, such as a cup wheel of an anemometer, should be identified by
number.  If they are not numbered, the operator should be asked to mark them
for identification.

                The audit report should contain a list of all the equipment
provided and used by the auditor, including model and serial numbers and time
of last calibration, where relevant.

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 39
                                                             Page:    2  of 20
                4.2,7.1.3  Where
                The audit form should have a space to show the location of the
sensor on a tower, including height.  A sketch is useful to show the  relative
positions of the sensing elements with respect to possible biasing influences,
such as the tower, other sensors and buildings.

                4.2.7.1.4 When

                The date or dates when the audit affected the system  operation
should be listed.  The time when the system or a particular sensor was taken
"off-line" and put back "on-line" should be listed.   The time, or time period,
when each datum value was taken is vital for the comparison with the  system
output.  Implicit in this is the need for the time the auditor uses to be
correlated to the time the operator or the system uses.  The auditor  should
rely on the National Bureau of Standards station WWV for correct time.  Battery
operated receivers, such as the Radio Shack Time Cube, are generally  available.

                4.2.7.1.5  How

                The audit form should either contain a copy of the method used
or reference the method number.  The audit report should contain copies of the
audit method used.  The methods should be detailed enough to identify each step
in the acquisition of the audit value and in the conversion of the value to
units compatible with the system output.

       4.2.7.2 Wind Speed

       There are two general philosophies in use by those who operate
anemometers in meteorological monitoring systems and networks.  The most coaaon
treats the system as a unit where the sensor and signal conditioner and
recorder are calibrated together.  The other, often employed by operators with
large numbers of anemometers, considers the sensor as a standard
interchangeable part.  In this case two audits are necessary.  One to challenge
the sensor calibration method and the other to challenge the system calibration
using a standard signal as a substitute for the sensor.  A full system audit
method from sensor input to system output can be used as a challenge  for the
system operated with interchangeable sensors.

                4.2.7.2.1 Sensor Control

                The controlled condition is rate of rotation of the anemometer
shaft.  The cup assembly or propeller is removed for this challenge.  The audit
form should provide space for fully defining the transfer function used by the
operators (usually supplied by the manufacturer).  This should include the
relationship of rate of rotation (R, rps) to wind speed (U, m/s), rate of
rotation to output volts (0, V) and rate of rotation to frequency (f, Hz), for
light chopper or a.c, generator types.  See 4.2.2.1.2 for a discussion of the
U=a+bR and U=bR types of transfer functions and how the constants "a" and "b"
are determined.  Some manufacturers provide the transfer function in  the fom
                             f = 26.439  (U - 0.281)
which can be converted to
                               U =  0.281 + 1.135R
once the number of pulses per revolution  (30 in this case) is known.

-------
                                                              Section No. 4-2.7
                                                              Revision No.    0
                                                              Date:    17 Sep 89
                                                              Page:     3  of 2Cp
                 The  rate  of  rotation can be  imposed
 on the  anemometer  shaft  in a number  of  ways.   If the
 method  is  to  drive the shaft with a  d.c.  motor,  the
 number  of  revolutions of  the shaft over a period of
 time  is the data value.   That value  divided  by the
 number  of  seconds  in the  time period gives average
 rate  of rotation R (rps).  The R is  converted to U
 by the  transfer  function  and the L) is compared to
 the system output  in the  same units  for exactly the
 same  period of time.   If  the system  provides 5
 minute  average speeds, the count is  for 5 minutes
 with  the start and stop  times inclusive of the
 system  period.   If the d.c,  motor is reasonably
 constant (± 10%),  a  few  seconds out  of
 synchronization  over 300  seconds is  acceptable.   The
 period  of  time,  however,  must be exactly 300 seconds
-which can  be  hand  timed with a sweep second  watch to
 about ± 0.2 s.   If the system only reports hourly
 averages,  and cannot be  changed to a shorter time,
 samples of the signal conditioner output voltage may
 be used to estimate  the  system output.   Three rates
 of rotation in addition  to zero are  recommended.
 Since the  important  speeds are low speeds and not
 full  scale speeds, the use of simulated speeds on
 the order  of  2,  5  and  10  m/s is acceptable.   Using
 the transfer  function above,  these speeds are
 simulated  by  R values of  1.51,  4.16  and 8.56 rps
 {90.6,  249.6  and 513,6 rpm).   Figure 4,2.7.1 is an
 experimental  d.c.  motor drive used for  this  kind of
 audit.   Figure 4.2.7.2 is a  second generation d.c.
 motor drive capable  of being powered by a D  cell and
 a  9 volt transistor  battery.
           If  a propeller  anemometer  with a transfer
 function of
                        U =  0.294R
Figure 4.2.7.2
 is  challenged,  the  speeds  of  2,  5 and  10  m/s  will be  simulated by 6.9,  17, and
 34  rps  (414,  1020,and  2040 rpm).   If the  auditor could  generate five R values,
 1.5,  4,  7,  17 and 35 rps,  both cups and propellers could  be challenged at three
 meaningful  speeds plus zero.   See Figure  4.2.7.3 for  a  third d.c.  motor system.
    c t
 Figure 4.2.7.3   An  experimental  12-volt  d.c,  motor and  counter

-------
                                                              Section No.  4.2.7
                                                              Revision No.     D
                                                              Date:    17 Sep H9
                                                              Page:     4  of 23
Climatranics  cup
Teledyne-Geotech  cap
R.M. Young propeller
  Figure 4.2.7.4
     A simple d.c.  motor might be made  to turn
the shaft, but the key to the audit challenge is
the measurement of the shaft revolutions.  A
light chopper and counter Is a straight forward
approach to this measurement.  Hand switching
the counter for periods as short as 60  s will
produce better than one percent accuracy in
time.  If the light chopper produced 10 counts
per revolution,  the count rate required for the
four R values mentioned above is 5 to 60 Hz, an
easy range for simple battery operated  counters.
A system such as described above has the
advantage of independence front commercial power,
a significant advantage for some remote wind
systems.

     The controlled condition audit requires a
hands-on policy regarding the sensor.   The
measurements of the starting torque does not
require the sensor to be connected to the
circuit.   It is possible, with proper equipment
and care, to install the speed challenge motor
on the sensor and operate that challenge with it
connected in its operating location.  Another
way to challenge an anemometer is to connect the
shaft to a synchronous motor.  The assumption
here is that the motor is running in
synchronization and the R value is therefore
known from the specifications of the motor or an
independent measurement of its rate of  rotation.
Figure 4.2.7.4 shows three anemometers  coupled
to a synchronous motor through a universal
coupler.

     The last measurement is the time constant.
With the simulated speed on its highest rate,
and with a meter on the output voltage of the
signal conditioner, turn off the d.c.notor and
measure the time it takes to reach the value of
the simulated speed minus 63% of the simulated
speed.   Examples of wind speed audit procedures
and forms are found in Figures 4.2.7.5  through
4.2.7.8.

-------
                                                        Section No.  4.2.7
                                                        Revision No.     0
                                                        Date:   17 Sep  89
                                                        Page;    5  of  20
            Cup Anemometer - MSI  method  CAOO3 (version  B/l/84)

       This method provides for a comparison  of  the transfer  -function
       used Hith the system to the output of  the system.   This  is done
       by causing the anemometer  shaft to turn at a known  rate  of
       rotation and observing the output.   The means of turning the
       shaft and measuring the rate of rotation  are provided  by the
       auditor and are completely independent of tRe operating  system.
       The method does not challenge the transfer function.   This can be
       done best with a wind tunnel test.

       The report form for this method includes  space for  an  optional
       determination of starting  torque and system time constant.  The
       torque measurement may be  used as an indication of  bearing
       condition and hence starting threshold of the anemometer.  The
       time constant is of use if  turbulence  is  measured.

       CAOO3-A  Remove the cup assembly.  Mount  a coupler  to  the
       anemometer shaft.  A i/B"  shaft is required.  If the anemometer
       does not use that size or  it is not  accessible, an  interface
       fitting will be required.   Clamp the drive motor to the  support
       column of the shaft so that the coupler is engaged  with  the drive
       wheel.  Determine if the cup assembly  turns the 'shaft  in a clock-
       wise or counter clockwise  direction, when viewed from  above.
       Clockwise is common and is used on the form.  Operate  the drive
       motor at two speeds (find  the desired  rps from the  transfer
       function) which are important to the application of the  wind
       speed data.   Use a time period synchronous with the system
       output.  An average of one minute or longer is required.  If the
       system provides only instantaneous samples of output volts, take
       12 samples over a two minute period  and use the average  of the
       samples to compare with the average  rate  of rotation measured.

       CA003-B  This method requires that the system be 'operating with
       all cables in place 
-------
                                                          Section No.  4.2,7
                                                          Revision No.    0
                                                          Date:    17 Sep B9
                                                          Page;     6  of 20
                  PERFORMANCE AUDIT REPORT by	      CAOO'

       MEASUREMENT SYSTEM - Cup anemometer
             System number 	a_	____	________________	____
             S en sor        ___ ________________________	___	_
             Cup  assembly  ,	_______	,	'	
             Loc a t i on       ____	 _______	:_._„	;	_____
             Signal conditioner	_	_______	.__	
             Data Channel	.	__„
       DATE 	/	/	TIME o-ff line	on  line	test start

       TRANSFER FUNCTIONS  (rps to mps) 	
                            Crpm to volts)	.	
                            pulses per revolution  	
       TEST RESULTS

         CAOQ3-A          — .challenge speed —    output   di-f-ference

                       time   revs,  rps    nips    mps    mps     7.

            0        	     0     O	,	

           "d.c."

            S CW     	,	.	

            F cia	,		,	


            Torques  _______  Oz.—In. cw.  Time constant	seconds

         CA003—B               expected    observed       dif-f erence
                      tiflie        mps         nips         mps     %

            0         	;	      	         	 	 	
Figure 4.2.7.6  Audit form for  the cup anemometer method

-------
                                                     Section No. 4.2.7
                                                     Revision Mo.    0
                                                     Date;   17 Sep 89
                                                     Page:    7  of 20
     Fixed Axis Propeller  -  MSI  method FAP001  (version  8/1/84)

This method provides -for a comparison  of  the trans-fer -Function
used with the system to the  output  o-f  the system.   A separate
form is provided for W  (vertical  component) since  a different
transfer function is o-ften used  -for this  direction than Is used
for U and V.  The method causes  the propeller  shaft to  turn at a
known rate of rotation while observing the output.  The means of
turning the shaft and measuring  the rate  of rotation are provided
by the auditor and are completely independent  of the operating
system.  The method does not challenge the transfer function,
This can tie done best with a wind tunnel  test.  The sign
convention used with respect to  clockwise and  counter clockwise
is that of the system being  challanged.   Differences are always
calculated by subtracting  the audit challange  value from the
system output.  Arithmetic convention  is  followed  even  though the
minus sign is used as an indicator  of  direction.   For example.
the difference between a -1,5 mps audit challenge  and a -1.3 mps
system output is •*•<>.2 mps  even though  the system underestimated
the speed  (a negative error) with respect to the audit  value in
the "-" direction.

The report form for this method  includes  space for an optional
determination of starting  torque and system time constant.  The
torque measurement may be  used as an indication of  bearing
condition and hence starting threshold of the  propeller.  The
time constant is of use if turbulence  is  measured.

FAPOOt—A  Remove the propeller.   Mount a  coupler to the propeller
shaft.  A  1/8" shaft is required.   If  the propeller does not use
that size or it is not accessible,  an  interface fitting will be
required.  Clamp the drive motor to the support column  of the
shaft so that the coupler  is engaged with the  drive wheel.
Operate the drive motor in both  a clockwise and counter clockwise
direction, when viewed from  in front of the propeller.   Operate
the drive motor at two speeds 
-------
                                                         Section No, 4.2.7
                                                         Revision No.     0
                                                         Date:   17 Sep 89
                                                         Page:    8  of 20
              PERFORMANCE AUDIT REPORT by	,	    FAPOOiW

    MEASUREMENT SYSTEM - Fixed axis propeller
         System number	:	_	
         Sensor         -,r -.„ ,,,t, ,„ „. _ u, •—_» __ ^^  ., __.^_ ^^ „ _—-^	,	-^ -- -...	.
         Pr ope 11 er      —_««.___^.^_;	,	;	—	
         Location	,	
         Signal conditioner'
         Data channel   	'
    DATE 	/__/,	 TIME o-ff line 	,	 on  line	teat start"	

    TRANSFER  FUNCTIONS 1 rps = 0.294 mps  <3  pulses per revolution)
                          CWCvolt-s>-2.33*4  » m/s

    TEST RESULTS

      FAPOOIW—A      — challenge speed —     output   difference

                    time   revs,  rps    mps     mps    mps     7.

         O	              .      0	,	,

       "d.c."

         s cw	.	

         s ccw	;_	



         F CCW  	,_	,	

       "sync"      time   rps    mps   volt     mps    mps      %

         S CW   	,  5.000   1.47	

         S CCW	S.OOO  -1.47	

         F CW   	  3O.OO   8. 32	.	

         F CCW	SO.00  -S.S2	,	,

         Torque:	Oz.-In. cw.	^.Os.-In.  ccw.  T.  Const.	s.

      FAPOOIW—B            expected    observed      difference
                   time        mps          mps        mps     %
         O

       test
Figure 4,2.7.8  Audit  form for the propeller anemometer method

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    9  of 20


                4.2.7.2.2  CTS Method

                The collocated transfer standard (CTS) method for wind speed
Involves mounting a carefully calibrated anemometer in the vicinity of the
subject anemometer being audited.  The CTS should have certificates tracing its
calibration to NBS or some other standard facility.  If the ASTM  (1984) method
for comparability is being used, the CTS needs to be within 10 m of the subject
aneaoraeter in the horizontal and the lesser of 1 m or H/10, where H is the
height above ground in aeters, in the vertical.  It is important to site the
CTS to be representative of the flow at the subject anemometer.  Mutual
interference should be minimized through siting and through editing out data
where the direction shows the wind-passing through one to reach the other.  The
accuracy potential of the CTS method is based on data taken in 1982 at the
Boulder Atmospheric Observatory  (1AO) and published by Finkelstein et al.
(1986) and Lockhart (1988).  The anemometers for this study were spaced about 5
m apart.   The closer together they are in the horizontal the larger the
direction sector of mutual interference.

                The best situation for CTS auditing is one in which both
aneaometers are connected to the auditor's data logger.   The element of the CTS
audit is the difference in speed calculated by subtracting the CTS speed from
the subject speed.  The method requires a sufficient number of simultaneous and
independent differences.  A simultaneous difference is one where the time
between sampling each anemometer output is less than O.lx, where T (s) is found
by dividing the distance constant, D (m), by the wind speed U (m/s).
Independence is achieved when the time between sampled pairs is larger than 4x.
For example, assume the subject anemometer has a distance constant of 5 m and
the CTS has a distance constant of 1 m.  If the wind speed is about 3 m/s, t
will be 5/3 = 1.7 s for the subject and 1/3 - 0.3 s for the CTS.  Simultaneous
samples will exist when the sampling rate of the data logger is less than the
shortest O.lT or 0.03 s in this example.  Most data loggers are fast enough for
the example.

                Independence is achieved when the time between successive
saaple pairs is long enough.  In the example above, 4r is 20/3 = 6.7 s for the
subject anemometer and 4/3 = 1.3 s for the CTS.  If the CTS logger is set for
one sample every 10 s, the data will be independent at 3 m/s.  The ASTM method
defines minimum sample size in terms of the resolution of the measured or
reported speed (assume 0.1 m/s for the example, a recommended resolution) and
the standard deviation of the series of differences.  It takes 900 times the
variance of the series to provide the minimum number.  If the two sensors are
well sited and properly operating, the variance will be small.  The BAD data
showed the standard deviation of the difference to be less than 0.2 m/s
(variance of 0.04).  This should be the minimum condition for a good CTS data
set.  If the variance is 0.04, the minimum sample size is 900 x 0.04 = 36.

                Assume the subject anemometer produces a scalar average speed
every 15 minutes and it is not possible to wire the output into the auditor's
data logger.  The CTS uses 90 samples, one each 10 s, to assemble its
concurrent 15 minute scalar average.  At this point, one can take one of two
paths.  One is to assume that the subject is operating well,  has a short enough
distance constant and is likely  to agree well with the CTS on a sample to
sasple basis; not perfectly because a large bias can still have a small
variance.   There is no way to verify this assumption unless the audit results

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   10  of 20


show good agreement.  Based on this assumption, each 15 minute value is a data
point with a sufficient sample size to compare the anemometers.  Each datura
point can be used in a linear regression analysis to define the subject
anemometer's accuracy as a function of wind speed, if enough dynamic range
exists in the data period.  In just a few hours,  then, the CIS method with this
assumption will produce a measure of accuracy.

                The other way is to treat each 15-minute average as a single
sample of 15-mlnute averaged data.  A 9 hour period will provide the minimum
saaple size of 36.  This is an awkward period of time.  It takes some time to
install the CTS and many auditors work 8 hour days like other people.  There is
no such thing as too many samples.  Added time usually enlarges the dynamic
range for the audit.  The optimum CTS audit goes for something like 24 hours,
one diurnal cycle.  For an example of this CTS audit, look at the data  listed
in Table 4.2.7.1.

                Each value is a 20-ainute scalar average.  The subject
ane»ometer is a Climatronics F460 cup (C-V-W).  The CTS is a Young Propeller
Vane (P-V-W) located 5 m away.  Both are at a height of 10 m.  The standard
deviation of the CTS sensor is shown for each 20-minute period.  The difference
between the subject anemometer and the CTS is shown in the column headed by
"Y-X."  Notice that the average difference is a small -0.12 m/s and the
standard deviation of the difference series is 0.10, half of the maximum
criteria.  To express the accuracy of the subject anemometer with respect  to
the CTS, a linear regression was run.  The constant of 0.02 m/s says there is
no bias of significance.  The X Coefficient of 0.96 says that there is  a 4 %
underestimate of speed at all speeds.  The best fit straight line through  the Y
points is calculated by multiplying the X (true) value by the coefficient and
adding the constant.  The re'sidual error is then found by subtracting the X
value from the estimated Y value, Y', {column headed by Y'-X).  Notice  that the
average difference  is 0.000, as it must be, and the standard deviation  is
lower, O.O7 m/s.  The audit report for this subject anemometer would report the
error of 4 % in  the slope of the subject transfer function.   (NOTICE: This is
an example.  The CTS is arbitrarily selected.  It is possible that in this case
the CTS was 4 % high or they were each off 2 '/,,  That is not important  for this
example and the  speeds are so close that this analysis was not included  in the
BAO experiment.)  These data are shown graphically in Figure 4.2.7.9 and Figure
4.2.7.10.  The anemometer data are shown on an XY plot with the best fit
straight line through the 72 points.  Figure 4.2.7.10 shows an XY plot  of  the
residuals.  Also shown is the normalized turbulence, 
-------
                                           Section No. 4.2.7
                                           Revision No.    0
                                           Date:   17 Sep 89
                                           Page:   11  of 20
Table 4.2,7.1 Simulated Audit Data (BAO 1982)
f/i2
Site/
UK

4 im
I 1820
4 1540
4 im
4 1120
4 1140
4I2SO
4 1220
4 1240
4 1530
4 1320
4 1340
4 IW
4 1420
4 1444
4 !S»
4 520
4 1540
4 im
4 !£20
4 IMO
4 1730
4 1T20
4 !?W
4 :wo
* IB20
4 1S40
4 !*)0
4 I!20
« *;r*0
4 300
4 2HO
4 2040
4 2100
4 2120
4 21 40
4 r-oo
4 IZO
4 22W
4 11-30
4 ins
4 :i40
5 0
5 23
5 «>
5 100
Y
Subject
Speed
il/5)
3.701
2.972
3.557
2.660
2.463
2.376
3.210
5.511
4.546
4.261
4.157
3.127
2.364
2.451
4.176
8.281
7.678
6.371
5,569
3.346
8.338
7.236
5,425
3.989
4.399
4.407
3.347
3.94!
4.344
4.300
2.907
2.861
2.049
1.747
3.796
4.065
3.S35
4.221
4.097
4.037
3.952
3.559
3.449
3.17!
3.840
4.393
I
CIS
Sseefl
(l/S>
3.729
2.979
3.555
2.605
2.365
2.399
3.344
5.623
4.722
4 cecr
• -JJ
4.455
3.250
2.433
2.498
4.270
8.601
7.B93
6.600
2.976
3.413
8.693
7.436
5.589
4.111
4.sl7
*.S15
4,012
4.121
4.474
4.474
3.035
3.156
2.221
2.015
3.957
4.198
4.031
4.376
4.261
4.194
4.135
3.717
3.530
3.290
4.040
4.617
s
CIS
Sipa
d/s)
1.047
2.441
2.7SO
0.844
1.412
0.883
1.951
3.230
1.115
1.319
1.348
1.120
1.557
1.039
5. 887
2.564
1.638
1.638
2.847
3.537
4.531
1.508
0.751
0.233
0.820
0.930
0.345
0.291
0.151
0.363
0.442
0.154
0,022
0.190
0.246
0.215
0.137
0.206
0.282
0.194
0.144
0. 109
0.123
0.035
0.19?
0.113
5/1
li/s)
0.23
0.82'
0.7S
0.32
0.60
0.37
0.55
0.57
0.24
0.29
0.30
0.34
0.64
0.42
1.33
0.30
0.21
0.26
0.43
1.04
0.52
0.20
0.13
0.04
0.18
0.20
0.09
0.07
0.03
0.0S
O.S5
0.05
0.01
0.09
0.06
0.05
0.03
0.05
0.07
0.05
0.03
O.C3
0.03
0.01
0.05
0.02
Y-X
(6/5)
-0.03
-0.01
0.00
0.05
0.10
-0.02
-0.13
-0.11
-0.18
-0.29
-0.30
-0,12
-0.07
-0.05
-0.09
-0.32
-0.21
-0.23
-0.41
-0.07
-0.36
-0.25
-0.16
-0.12
-0.22
-0.21
-0.16
-0.18
-0.13
-0.17
-0.13
-0.29
-0.17
-0.27
-0.16
-0.13
-0.15
-0.16
-0.16
-0.16
-0.18
-0,16
-0.13
-0.11
-0.20
-0.22
'f-1
U/s>
-0.11
-0.10
-0.13
-0.14
-0.17
-0.06
0.02
-0.10
-0.00
0.12
0.13
O.Oi
-0.01
-0.04
-0.06
-0.02
-0.10
-0.03
0.13
-0.05
0,01
-0.04
-0.05
-0.03
0.05
0.04
0.02
0.03
-«,04
0.01
•5.02
0.18
0.10
0.21
0.02
-0.02
-0.00
-C.Ol
0.01
0.00
0.03
0.02
0.00
-0.01
0,05
0.05


5
c
J
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
c
»j





120 3.822 3.955 0.881
140 2.198 2.219 0.520
200 1.630 1.742 0.059
220 1.484 1.510 0,576
240 i.599 1.624 0.377
300 2,569 2.697 0.076
320 1.662 1.696 0.341
340 1.769 1.934 0.190
400 1.054 1.114 0.459
420 1.471 1.455 1.217
440 2.181 2.286 0.438
500 1.338 1.419 0.680
520 3.315 3.501 0.180
540 3.060 3.212 0.220
600 1.992 2.074 0.379
620 1.765 1.836 0.336
640 0.920 0.896 0.072
700 0.636 0.523 0.280
720 1.809 1.856 0.422
740 1.571 1.577 0.173
800 2.05B 2.115 0.292
820 1.910 2,000 0.519
340 1.181 1.165 0.345
900 2.880 3.013 0.891
920 4.2S6 4.402 0.537
940 3.839 4.000 1.752
Average
Standard deviation



0.22
0.23
0.03
0.38
0.23
0.03
0.20
0.10
0.41
0.34
0.19
0.43
0.05
0.07
0,18
0.18
0.08
0.54
0.23
0.11
0.14
0.24
0.30
0.30
0.12
0.44





-0.13
-0.02
-0.06
-0.03
-0.02
-0.13
-0.03
-0.17
-0.06
0.02
-0.10
-0.08
-0.19
-0.15
-O.Oi
-0.07
0.02
0.11
-0.05
-O.Oi
-0.06
-0.09
0.02
-0.13
-0.12
-0.16
-0.125
0.100



-0.01
-0.05
0.01
-0.01
-0.02
0.04
-0.01
0.11
0.04
-0,05
0.03
0.04
0.06
0.04
0.02
0.02
-0.04
-0.11
-0.01
-0.04
-0.01
0.03
-0.04
0.03
-0.05
0.01
0.000
0.069

Regreision Output:
















Constant
Std Err of Y Est
R Squared
No. of Observations
Degrees of Freedoi

X Coeff. 0.9576
Std Err 0.0048








































0.0233
0.0699
0.9982
72
70












-------
                                                       Section No. 4.2.7
                                                       Revision No.    0

                   SIMULATED CTS AUDIT      ?£:   II s?f *o
             Propeller Vane (P-V-W)  as Collocated Transfer Standard
          Regression Analysis
          Y = 0.023 + 0.96 X
                                            Data: Boulder Experiment
                                                 9/4-5/1982
                                            72 — 20 minute averages
                    CTS (P-V-W)  Wind Speed  (m/s)
Figure 4.2.7.9  XV plot of simulated wind speed audit data
 CO

 E

 "§  0.2
 o

 
                                                                    c/5
                                                                 o.
                                                                10

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   13  of 20
                4.2.7.2.3  "W Anemometers

                Some stations measure the vertical component of the wind, with
an anemometer sensitive only to the vertical component of the wind.  A
vertically »ounted helicoid propeller, or "W" propeller, is the most common
instrument for this measurement.  The same audit methods can be used as are
used on a propeller anemometer.  A synchronous or d.c. motor will challenge the
rate of rotation vs. wind speed and a torque device can be used to find the
starting threshold.  The common manufacturer's recommended practice is to use a
different transfer function for the W propeller than the one applied when the
same propeller is used for vane-oriented speed or for the N-S and E-W
components of a UVW anemometer.

            4.2.7.3 Wind Direction
                4.2.7.3.1 Sensor control

                The first thing to do on a performance audit of a direction
vane is to record the as-found orientation value.   Have the operator hold the
vane so that it points to or from (whichever is most accurate for aiming) the
distant orientation target.  Verify the alignment by viewing the vane and
target from the ground.  Move back away from the tower or mast on a plane which
passes through the sensor and the target and verify that the vane is in the
plane also.  Field glasses or a theodolite can help make this sometimes
difficult observation.  The vane must be held steady or clamped until a
constant output exists for a few minutes.  Record this value.

                The colntrolled condition for a wind vane is a relative position
of the vane with respect to the sensor housing.  There are several ways to
impose a series of known relative positions on the vane-sensor combination.
They vary in effective accuracy.  It is critical to know the time constant of
the direction circuit BEFORE starting the performance audit.  It can be
measured by setting the vane to a known direction, simulate a wind from 090°
holding the vane steady until the 090  (or voltage equivalent) output is
steady.   Move the vane quickly (< 1 s) to 270  and measure the time constant of
the system.  Assume that a time constant of 3 s is measured.  Table 4.2.7.2
shows the change in output angle and voltage (assuming a 540  format and 5V
output)  as a function of time.

                     Table 4.2.7.2 - Time Constant Effects
Time
Constant
(No.)
0
0.2
1
2.3
3
4.6
6.9
Time
Angle
(sec.)
0
0.5
3
6.9
9
13.8
20.7
Vane
Angle
(deg.)
090
270
270
270
270
270
270

Angle
(deg. )
090
106
204
252
261
268
270
Error (54085)
(deg.) (volts)
0
164
66
18
9
2
0
{after Fritschen
O.S33
0.981
1.889
2.333
2.417
2.483
2.498
and Gay,
Change
( % )
0.0
9
63.2
90.0
95.0
99.0
99.9
1979)

-------
    Section No.  4.2.7
    Revision No.     0
    Date:   17 Sep  89
    Page:   14   of  20
              :-^^4?f'
                Notice that in this example a 180  shift requires waiting 20
seconds for the reading to be representative of the new position.  If a 90
shift is used, 14 seconds will provide an output within 1  of the final value.
If measurable time constants are found, suggest to the operator that the
lanufaeturer be called about steps which might be taken to modify the circuit to
a ainimui time suitable for 60 Hz noise filtering,

                The least accurate method for challenging the relative position
accuracy of a wind vane is to point the vane in various directions while still
mounted on the tower.   This can provide positions related to external objects
rather than constant angle changes.  It is estimated that the accuracy of this
nethod is two to five degrees, with the exception of a parallel alignment.  The
tail vane can be located parallel to a cross arm to within one degree, and held
parallel on a calm day.

                A second method puts the operating sensor in a controlled
situation like a room where the electronics are located.  The sensor can be
placed at the center of a template with radial lines every 60 ,   The sensor can
be oriented to the template and the vane
aoved and clamped when the vane is parallel
to the radial line.  If care is taken to        _','...       ;  •*
avoid parallax errors (non-parallel or        ""•'./
ion-perpendicular observations) this method     ><3^v
can provide relative accuracy on the order
cf one degree.                                '/?-'-"

                The best method replaces      ;: ,'.^,
:ie vane with a fixture with the capability    -T-;-V
cf holding the shaft in fixed positions
• ith respect to the sensor housing.           '"
Fixtures of this type can provide               C  ;
repeatable position accuracy of < 0.1 .
figure 4.2.7,11 shows such a device.  A
different application of this precise
nethod uses a theodolite base as the mount
err the sensor.  With the vane or vane
substitute held in one position, the base
c±n be rotated in very accurate steps.
Theodolite worm gear assemblies divide a
circle in whole degrees with a vernier
adjustment with 0.1 degree index marks far         -
T-ough apart to allow easy interpolation to
I 32 degrees, a resolution wasted on the
Application of wind direction measurement.

                The audit report form
s.iould contain the transfer function used
:: convert output voltage to azimuth
Agrees.   This may include a 540 format
-~.ere azimuth values greater than 360 are
reduced by subtracting 360.   The report
ftru should also contain the challenge
rr-gression used by the selected method.                 Figure 4.2.7.12
Figure 4.2.7.11
   &IV-

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   15  of 20
For example, Fig. 4.2.7.14 and Figure 4.2,7.15. show an audit method and audit
form for wind direction which specifies 16 relative angles, each 60  from the
last one,  in a clockwise rotation for 420  followed by a counterclockwise
rotation of 480  .  This tests a 540 format, provides four samples at 180
(duplicates from each direction) and a duplicate counterclockwise 240  pair.
The report of this series describes the range of relative error resulting from
the shaft position measurement of the sensor {see 4.2,2.2.2.3 for an example).

                The starting threshold of the bearing and transducer assembly
should be measured by some method (see 4.2.2.2.1.2 and 4,2.2.2.1.3).  If the k
value is not available to convert the torque to threshold speed at some
accuracy angle, the operator should be requested to ask the manufacturer to
provide it for the next audit.

                The bearing to the orientation target should be independently
challenged with a method capable of better than compass accuracy.  A theodolite
is ideal for finding the bearing to other distant objects,   A solar observation
is recommended (see 4.2.4.3.2),

                The last activity of the sensor control audit is to repeat the
orientation test described above for the as-found value.  The as-left value
will represent any changes the operator may have made and the new orientation,
if the sensor was not keyed for orientation.

                4.2.7.3.2 CTS Method
                There is no technical need for a CTS audit of direction.  No
new information is added by this method to that gained in the sensor control
method.   As a parallel example to the simulated CTS speed audit, data from the
sa»e period of time from BAG is shown here, structured as a simulated CTS
direction audit.   Table 4.2.7.3, sorted for ascending CTS direction, shows the
2Q-»inute average direction for both the CTS (Young Propeller Vane, P-V-W) and
the subject (Climatronics F460, C-V-W).  Also shown is the 
-------
                                                  Section No.  4.2.7
                                                  Revision No.     0
                                                  Date:   17  Sep 89
                                                  Page:   16   of 20
Table 4.2.7.3  Simulated CTS Direction Audit Data
«/82
D'ttTff
TiM

5
3
5
5
4
4
4
5
4
4
4
4
5
4
4
4
X
4
5
5
4
4
5
5
5
4
5
3
5
4
4
4
5
3
5
5
5
4
4
3
5
5
5
5
5
4
4
5
4
4
5
•*
5
5
t
t,
t
5
l.
4
4
3
*
5
-1
•i
4
5
1
*
4
*

f


740
72O
1140
44O
1000
1840
182O
1O40
220O
2120
2140
190O
1100
1620
172O
1700
174O
1800
1020
620
1640
1920
500
20
0
2320
112O
too
4O
230O
222O
2340

39.34
20.76
87.87
6.50
27.02
10.17
30.55
16*28
7.62
4.04
10.87
9.OO
13.26
6O.97
6.10
8.98
6.66
17.58
27. O4
32.37
9.24
9.10
30.40
4.44
5.99
6.94
22. B4
3.45
5.35
6.21
7.9O
5.88
11.83
9.57
9.31
8.00
20.18
S.B6
8.76
6.29
17.04
9.38
28.91
9. BO
25.28
35.85
42. 3O
63.08
29. 3O
12.10
35.26
11.43
20.16
12.57
23.11
29.16
22.09
43.63
34.09
21.27
27.46
29.47
11.74
65.04
22.86
17.77
10.46
21.03
12.26
67, 66
30. 58
20.56
Y-X



-------
                      SIMULATED  CTS AUDIT
              Propeller Vane (P—V-W) as Collocated Transfer Standard
                                                           Section No. 4.2.7
                                                           Revision No,    0
                                                           Date:   17 Sep 89
                                                           Page:   17  of 20
360 -
Regression Analysis
Y =  -0.22 +  1.005 X
Data: Boulder Experiment
     9/4-5/82
72 — 20 minute overages
      180
   0)
   TJ

   c
   ,2
   "5
  5  27°
  ~o
   c
  >
   !
  u
  -t->
   o
   3
  to
                       090           180           270           360

                CTS  (P-V-W) Wind Direction [Theta]  (deg.)

Figure 4.2.7.13  Simulated CTS Vind Direction Audit  Data.

                4.2.7.3.3   Vertical Wind Direction,  ^

                The sensor control method is used for bivane auditing.   The
vertical part of the  bivane operation is treated in  the same way as the
horizontal part, except  different fixtures are used.  Special fixtures are
required for each bivane design, but the principle is the same.   A relative
zero point is set when the fixture is attached to the bivane.  That point is
where the vane shaft  is  perpendicular to the vertical axis of the sensor.  From
this starting point where  the output should be the equivalent of 0 ,  the vane
is held in 15  steps  until its physical limits are reached, both tail up ( + )
and tail down (-).  Threshold is very hard to measure on a bivane because of
the static balance conditions of the vane.  If the vane is perfectly balanced
and its remains where ever it is physically moved, a force gage measurement at
soae distance from its axis of rotation will yield the starting threshold just
as the vane begins to move.

       4.2.7.4 TURBULENCE  OR 
-------
                                                             Section No.  4.2.7
                                                             Revision No,     0
                                                             Date:    17 Sep 89
                                                             Page:    18  of 20


       The challenge should be realistic or at  least  within some realistic
range.   The challenge must take into consideration the wave shape of the
variable direction imposed on the system in calculating the true slgma value
with which the output will be compared.   The effective time constant of the
direction system, calculated from the delay distance  of the sensor and some
nominal wind speed important to air pollution applications, should define the
        frequency used in the sigma challenge.

-------
                                                        Section No. 4.2,7
                                                        Revision No,    0
                                                        Date:   17 Sep 89
                                                        Page:   19 of 20
        Wind Vane  - MSI method WV004  {version 8/1/34)

   This  method  describes  the relative performance of the  wind vane
   as  &  shaft-position transducer and the orientation  of  the
   transducer with respect to true North.  The former  is,done with a
   fixture,  part of which is mounted to the transducer body and  part
   mounted  to the  shaft in place o-f the vane.  The  latter requires a
   determination of true  North  (see MSI method SN008)  and a setting
   of  the transducer relative to that orientation.

   The report form for this method includes space for  the optional
   method to define the "open space" where relevant to the sensor
   and application!  Also there.is space to record  starting torque
   measurements and system time constant estimates  for turbulence
   Measurement.

   WV004-A   Remove the wind vane assembly  
-------
                                                         Section No. 4.2,7
                                                         Revision No.    0
                                                         Date:    17 Sep 89
                                                         Page:    20  of 20
             PERFORMANCE AUDIT REPORT  by	__	„_„_      WV004

  MEASUREMENT SYSTEM - Wind vane
       System number	^ _„„_,„.,.,...„„„,„	_.	„___«,___«•«.«_«.__;	,
       S*n*or    ,   . ______i^______^,.._-_.__________	i	„
       Vane        ..  	,	= =7=7: 7=	__.	,	,	;	_,_' ..
       Location	'	_	._	._ '
       Signal conditioner
       Data channel •' '	
  DATE	/	/	 TIME off line	on line ______ test  start  	,	

  TRANSFER FUNCTION 
-------
                                                             Section No. 4.2,8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1


4.2.8  ESTIMATING ACCURACY AND PRECISION

       4.2.8.1 Measurements      •               .

       Section No. 4.1.5 contains a detailed discussion of methods of
estimating accuracy, precision and bias using wind speed and wind direction as
examples.   That material will not be repeated here.  The measurement process
begins with an instrument which has some element or part which is sensitive to
the variable of interest.  If interest is in air flow with respect to the
surface,  there are two variables, wind speed and wind direction.  Each part of
this section discusses various aspects of how the measurements might be made,
calibrated, operated, maintained and documented to support a claim of
measurement validity.  It is recognized that instruments are usually parts of
data systems with sampling, processing and summarizing routines designed to
produce final elements of a data base to be used for some application.  The
earlier parts of this section were devoted to methods of tracking the
measurement process all the way through the system to the system output.
Accuracy was addressed in terms of how well what was designed to be done was
actually done.  The second part of this sub-section will deal with how well the
system design serves the application.

       4.2.8.2 Sinaaarized Data

       Summarization schemes are many and preclude a full discussion here.  The
auditor should define the methods used and comment on the appropriateness of
the method to the application of the summarized data.  There may be concurrent
suanarizations such as a scalar wind speed, a resultant vector wind speed and
some kind of summarized wind direction.  The accuracy of the data system should
reflect estimated errors because of an inappropriate summarization program.

       A software analysis is required to be sure  that the declared method of
summarization is in fact being accomplished by the computer program.  For
example,  is the scalar average wind direction avoiding the error of averaging a
circular range with a discontinuity?  If the average of winds ranging between
30O  and 06O  turns out to be about 180 , this problem still exists.

-------
                                                             Section No. 4.2.9
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   1  of  3


4.2.9     REFERENCES

Acheson,  D.  T.,1970:  Response of cup and propeller ratios and wind direction
    vanes to turbulent wind fields. Meteor. Monogr., No,33, Amer. Meteor. Soc.,
    pp. 252-261.

Acheson,  D.  T. ,1988:  Comments on "Anemometer Performance Determined  by ASTM
    Methods." J.  Atmos. Oceanic Techno1., 5 ,pp. 381-382.

ASTT-1.1984:   Standard Practice for DETERMINING THE OPERATIONAL COMPARABILITY OF
    METEOROLOGICAL MEASUREMENTS, D 4430-84. Amer. Soc. for Testing and
    Materials, Philadelphia, PA 19103.

ASTM.1985a:   Standard Test Method for DETERMINING THE PERFORMANCE OF A CUP
    ANEMOMETER OR PROPELLER ANEMOMETER  (Draft 6 of D22.ll). Amer. Soc. for
    Testing and Materials, Philadelphia, PA 19103.

ASTM,1985b:   Standard Test Method for DETERMINING THE DYNAMIC PERFORMANCE OF A
    WIND VANE (Draft 8 of D22.ll)  Amer. Soc. for Testing and Materials,
    Philadelphia, PA 19103.

Blackadar,  A. F.,1985:  Almanac for a weather station. Heldref Publications,
    Washington,  DC 20016.

Baynton,  H.  W.,1976:  Errors in wind run estimates from rotational anemometers.
    Bull,  Aaer.  Meteor. Soc., 57, 1127-1130.

Box, G,  E.  P., W. G. Hunter and J. S. Hunter, 1978:  Statistics  for
    experimenters. John Wiley & Sons, ISBN 0-471-09315-7.

EPA, 1987a:   Ambient Monitoring Guidelines for Prevention of Significant
    Deterioration (PSD), EPA-450/4-87-007, Office of Air Quality Planning and
    Standards, Res.  Triangle Park, NC 27711.

EPA, 1987b:   On-Site Meteorological Program Guidance for Regulatory Modeling
    Applications, EPA-450/4-87-013, Office of Air Quality Planning and
    Standards, Res.  Triangle Park, NC 27711.

Finkelstein. P.  L.,1981:  Measuring the dynamic performance of wind vanes.
    J.  Appl. Meteor., 20, pp. 588-594.

Finkeistein, P.  L.,J. C. Kaimal, J. E.Gaynor, M. E.Graves and T.  J.Lockhart,
     1986:  Comparison of Wind Monitoring Systems. Part I:   In-Situ Sensors.
    J.  Atmos. and Oceanic Techno 1. , 3,  pp. 583-593,

Finkelstein, P.  L.,  J. C. Kaimal, J. E.Gaynor, M. E.Graves and T. J. Lockhart,
     1986:  Comparison of Wind Monitoring Systems. Part II:  Doppler Sodars.
    J.  Atmos. and Oceanic Technol. , 3,  pp. 594-604.

Fritschen,  L. J.  and L. W. Gay, 1979:   Environmental instrumentation.
    Springer-Verlag, N.Y. ISBN 0-07-033175-8.

-------
                                                             Section No.  4.2.9
                                                             Revision No.     0
                                                             Date;   17 Sep  89
                                                             Page:   2  of   3


Gill.  G.  C.,1967:  On the dynamic response of meteorological sensors and
    recorders. Proceedings of the First Canadian Conference on
    Microneteorology, Part 1. Meteorological Service of Canada,  Toronto.

Gill,  G.  C.,1973:  The Helicoid Anemometer. Atmosphere, 11, 4, pp.  145-155.

Hayashi,  T.,   1987:  Dynamic response of a anemometer. J. Atmos.  Oceanic
   Techno 1.,  4, pp. 281-287.

Hoehne,  M.  E.,1973:  Standardizing Functional Tests. IEEE Transactions on
    Geoscience Electronics, Vol GE-11, No. 2, April.

Huschke,  R. E.,1970;  Glossary of Meteorology. Amer. Meteor. Soc.,  Boston, MA
     02108

Kaiaal,  J.  C. and J, E. Gaynor, 1983:  The Boulder Atmospheric Observatory,
    J. Appl.  Meteor., 22, pp. 863-880.

Kaiaal,  J.  C., J. E. Gaynor, P. L. Finkelstein, M. E. Graves, and
    T. J. Lockhart, 1984:  A field comparison of in situ meteorological
    sensors.  NOAA/BAO Report Mo. Six.

Koodo, J. ,  G. Mai to, and Y. Fujlnawa,  1971:  Response of Cop Anemometer  In
    Turbulence. J. Meteor. Soc, of Japan, 49, pp.63-74.

Lockhart, T.  J., 1977:  Evaluation of  rotational anemometer errors. Bull. Amer.
    Meteor. Soc., 58, pp. 962-964.

Lockhart, T.  J. , 1978:  A field calibration strategy for rotating anemometers
    and wind vanes. Proceedings of the 4th Symposium on Meteorological
    Observations and Instrumentation,  Denver CO, April 10-14. pp.  57-60.

Lockhart, T.   J., 1985a:  Some cup anemometer testing methods. J.  Atmos.  Oceanic
    Technol., 2, pp. 680-683.

Lockhart, T.   J., 1985b:  Wind-Measurement Calibration. Bull. Amer.  Meteor,
    Soc., 66, p.1545.

Lockhart, T.   J. , 1987:  Performance of an anemometer determined  by the ASTM
    method. J. Atmos, Oceanic Technol., 4, pp. 160-169.

Lockhart, T  .J., 1989  Accuracy of the collocated transfer standard method for
    wind instrument auditing. J. Atmos, Oceanic Technol., 6, pp.  715-723. 6/88)

MacCready„  P. B.,Jr. and H. R. Jex,1964:  Response characteristics and
    meteorological utilization of propeller and vane wind sensors.  J. Appl.
    Meteor. ,   3,  pp. 182-193.

MacCready,  P. B.,Jr.,1965:  Dynamic response characteristics of  meteorological
    sensors.  Bull. Amer. Meteor. Soc., 46, 533-538.

MacCready,  P. B.Jr.,1966:  Wind speed  measurements in turbulence.  J, Appl.
    Meteor.,   5,  pp. 219-225.

-------
                                                              Section No,  4,2.9
                                                              Revision No,     0
                                                              Date:    17 Sep 89
                                                              Page:    3  of  3


Middleton,  W. E. K. and A. F. Spilhaus,  1953:  Meteorological Instruments,
    University of Toronto Press.

Natrella,  M. G,,1966:  Experimental Statistics.  National Bureau of Standards
    Handbook 91.

Snow, J.  T., D. E. Lund, M. D.Conner,  S.  B.  Barley and C.  B.  Pedigo,1989:   On
    the dynamic response of a wind measuring system.  J.  Atmos,  Oceanic
    Techno 1. , 6, pp. 140-146.

Stearns,  C. R,,1985:  Wind-Measurement Calibration,  Response. Bull.  Anrnr.
    Meteor. Soc., 66, p.1545.

Sutton, 0.  G., 1955:  Atmospheric Turbulence.

Turner, D.  B.,1986:  Comparison of three methods for calculating the standard
    deviation of the wind direction. J.  Climate  Appl.  Meteor.,  25,  pp. 703-707.

Wieringa,  J. ,  1967:  Evaluation  and design of wind vanes. J. Appl.  Meteor.
    6, pp.  1114-1122.

-------
                                                            Section No,  4.3.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    1  of  3


                                 Section 4.3
               QA FOR TEMPERATURE AND TEMPERATURE GRADIENTS (AT)
                                   OUTLINE

Section                                                    Pages  Rev.  Date
4.3.0 OUTLINE AND SUMMARY                                  3     0     9/89

4.3.1 TYPES OF INSTRUMENTS                                 4     0     9/89
       4.3.1.1  TRANSDUCERS
                4.3.1.1.1  Thermistors
                4.3.1.1.2  Wire Bobbins
                4.3.1.1.3  Mercury-in-glass Thermometers
                4.3.1.1.4 Themocouples
       4.3.1.2  SIGNAL CONDITIONING
       4,3.1.3  ASPIRATED RADIATION SHIELDS

4.3.2  SPECIFICATIONS                                       4     0     9/89
       4.3.2.1  DELTA TEMPERATURE (AT)
       4.3.2.2  TEMPERATURE

4.3.3  ACCEPTANCE TESTING                                   2     0     9/89

4.3.4  INSTALLATION                                         1     0     9/89

4.3.5  CALIBRATION                                          9     0     9/89

       4.3.5.1  EQUIPMENT AND METHODS
       4.3.5.2  DELTA TEMPERATURE (AT)
       4.3.5.3  TEMPERATURE
       4.3.5.4  CALIBRATION REPORT EXAMPLE

4.3.6  OPERATIONS, MAINTENANCE AND QUALITY CONTROL           3     0     9/89

       4.3.6.1  OPERATIONS
       4.3.6.2  MAINTENANCE
       4.3.6.3  QUALITY CONTROL

4.3.7  PERFORMANCE AUDIT METHODS                            8     0     9/89

       4.3.7.1  GENERAL CONSIDERATIONS
                4.3.7.1.1  Who
                4.3.7.1.2  What
                4.3.7.1.3  Where
                4.3.7.1.4  When
                4.3.7.1.5  How

       4.3.7.2  DELTA TEMPERATURE (AT)
                4.3.7.2.1  Sensor Control  Method
                4.3.7.2.2  CTS Method

-------
                                                            Section No. 4.3.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    2  of  3
Section                                                     Pages  Rev.  Date
      4.3.7.3  TEMPERATURE
                4.3.7.3.1  Sensor Control
                4.3.7.3.2  CTS Method

4.3.8 ESTIMATING ACCURACY AND PRECISION               '.,!,.. ,0      9/89

      4.3.8,1  MEASUREMENTS
      4.3.8.2  SUMMARIZED DATA

4.3.9   REFERENCES                                           10      9/89

-------
                                                             Section No. 4.3.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  3

                       QUALITY ASSURANCE FOR TEMPERATURE

                         AND TEMPERATURE GRADIENT (AT)
4.3.0  SUMMARY

    The measurement of temperature Is standardized in great detail by all those
organizations Interested in such procedures, ASTM, ISA (Instrument Society of
America),  anil IMS  (Temperature Measurement Society).   The problem with
meteorological applications is that the free air temperature is required.  This
means the transducer needs to be exposed to the atmosphere which is in turn
exposed to the sun about half the time and to the very cold outer space the
other half.  The atmosphere is conditionally transparent to heat sources (the
sun) and sinks (outer space) so shielding must accommodate a wide range of
radiative conditions.  Wind also Influences the temperature shield.  It
transfers heat to  and from the shield in a variable way as a function of wind
speed.  Most effective shields use forced aspiration to expose the transducer
to nearly unmodified outside air.  Wind speed may also play a variable role in
the performance of the aspiration system.

    This section concentrates on the meteorological applications of air
temperature Measurement and the differential temperature measurements which are
interpreted as temperature gradients and applied as a measure of vertical
stability.

    Since the application of the measurement should define the accuracy needed,
both the relatively course air temperature and the relatively fine temperature
difference measurements will be considered.

-------
                                                             Section No. 4.3.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  4
4,3.1  TYPES OF INSTRUMENTS
    There are several materials and structures which change in some way as a
function of temperature.  General books on meteorological instruments such as
Mason and Moses  (1984), Mlddleton and Spilhaus (1953) and particularly Brock
and Nicolaidls  (1984) will provide details on a variety of these sensors.  From
the standpoint of quality assurance, a few basic principles and a few standard
types will represent the vast majority of instruments in use for air quality
applications.

    The measurement of temperature for air quality applications is generally
thought of as either air temperature, T, or a difference between two
temperature measurements, AT.  The application of these different measurements
require different specifications and auditing methods.  The two types of
temperature measurement will be treated in later sections as separate
measurements.

    Temperature  instruments are made up of three important parts.  The
transducer is the device which changes its electric value as a function of the
te»perature of  the transducer element.  The signal conditioner and cables
convert the electric value to a recordable output, usually volts.  The
aspirated radiation shield is the mounting structure which holds the transducer
in the atmosphere where the temperature is to be monitored.  Each of these
three parts will be discussed separately since there are various combinations
possible.

       4.3.1.1   Transducers

       Consider  the transducer as the part containing the sensing element.  In
most cases, the  sensing element is the transducer in air quality monitoring
applications.  The element is usually a thermistor (or thermistor network) or a
winding of fine  wire on an insulated bobbin.  It could also be a thermocouple
or a circuit element like an integrated circuit  (see Cole, 1978).  The elements
are usually encased in a protective capsule and sealed.  From an operational
standpoint it only matters how the sensor reacts to the temperature inside the
aspirated radiation shield.

                 4.3.1.1.1  Thermistors

                 The thermistor is an electronic semiconductor made from certain
metallic oxides, such as nickel, manganese, iron, cobalt, copper, magnesium,
titanium and other metals.  It is a nonlinear element.  One common supplier
(Yeilow Springs  Instrument Co. [YSI]) sells both the standard thermistor and
the "linear" thermistor.  Table 4.3.1.1 shows a typical negative thermal
response curves  of raw  thermistors and the nearly linear response of the
network thermistor.  Also shown is the positive response of two platinum RTD
(resistance temperature detector or resistance thermal device) for contrast.
Notice the large average change per  C with the YSI bead thermistor between
10°C and 2Q°C  (2220) as compared to the network thermistor (126Q) or the 100Q
RTD (Q.4Q) or  the 10000 RTD  (3.SO).  The raw bead thermistors are included
because, in the  future, microprocessor-based data systems can handle nonlinear
transducers as  easily as linearized ones.  The "linearized" YSI has a small
oscillating error of about ± 0.1°C with a wave length of about 4o"c.  The
impact of this  error on temperature difference systems is small.  At the

-------
                                                             Section No.  4,3.1
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  4
               Table 4.3.1.1 - Sensor Resistance vs.Temperature


T
('«
-10

0

10

20

30

40

50




Thermistors
y_ y_-
Ybi ibl
44031
(ohms) (AQ)
16600
-684.0
9796
-382.5
5971
-222.3
3748
-133.1
2417
-81.9
1598
-51.7
1081
44203 network
(ohms) (AQ)
13438
-127.9
12159
-125.3
10906
-126.1
9645
-128.6
8359
-128.8
7072
-123.6
5836
Platinum RTDs
_-.—_WTMPn____ UV fit

5-100 Pt
(ohms) (AQ)
96.09
0.391
100.00
0.390
103.90
0.389
107.79
0.388
111.67
0.387
115.54
0.385
119.39

1000Q Pt
(ohms) (AQ)
961 . 84
3.816
1000.00
3.804
1038.04
3.792
1075.96
3.780
1113.76
3.768
1151.44
3.756
1189.00
(AQ) is ohms per degree C for the 10 degree range
YSI 	 Yellow Springs, OH 45387
MINCO 	 7300 Commerce Lane, Minneapolis, MN 55432
HI-CAL 	 9650 Telstar Ave. El Monte, CA 91731-3093
steepest slope it is 0.025 C perodegree difference.  A AT would need to be ±
4°C before the error reaches 0.1°C, at which point the error is moot with
respect to application.

                The big advantage to thermistors is the relatively large
resistance of the element with respect to the resistance of the signal cable.
Lockhart and. Gannon, 1978, pointed out the fact that it would take a 12.50
difference in cable resistance to two sensors of a AT pair to cause a 0.1 C
error (bias) in the AT measurement.  Most signal conditioning circuits have the
capability of adjustment to eliminate such a bias.  Another little known
advantage is stability.  A several year stability test conducted at NBS showed
thermistors to be extremely stable.  This is contrary to early experience which
suggested that thermistors often failed with a shift in the transfer function
(ohms vs. temperature) gaining them a reputation of instability.  Better
packaging designs and better handling practices have eliminated many of these
problems.

                4.2.1.1.2  ¥ire bobbins

                The resistance of a wire changes with temperature.  If a long
piece of fine wire can be handled in some way, it can be used as a temperature
element.  Winding the wire on a non-conductive bobbin is the traditional method
of handling.  When the bobbin is large and open, a very fast response sensor is
made.  When the bobbin is potted in a stainless steel jacket, a more
traditional slower response sensor is made.  The transducer in both cases is a
length of wire.  Different metals have different temperature coefficients.  A

-------
                                                             Section No. 4.3.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  4

                                      0
40 gauge (0.08 mm diameter) wire at 20 C made of annealed copper has a
resistance of 3.4Q/m.  If it were made of German silver the resistance would be
65.90/m.  Platinum and iron are each 20.0 Q/m while nickel is 15.6 Q/m.
Manufacturers, choosing for stability, ease of handling and cost for a suitable
resistance,  have settled on a few materials.  Platinum (100Q) is the most
common for meteorological applications.  Nickel-iron is another common wire
providing a higher resistance at a lower cost with good stability.

                Because of the small resistance change for a 1 C temperature
change {0.4Q3, the transducer resistance must be measured with both high
resolution and attention to cable resistance.  Three and four wire bridge
circuits are commonly used, the latter being best for handling long cables.  It
will be shown in the sections on calibration and auditing that many of the
details of the transducer and circuits need not be known.  Only the system
performance is important.

                4.3.1.1.3  Mereury-in-glass thermometers

                These thermometers are not transducers, but they are commonly
used for calibrations.  Some styles have enough resolution to be read to 0.02°C
with some care.  The ASTM series of Precision thermometers are examples of
these.  They are 37.9 cm (15 inches) long and breakable (and expensive - $50 to
$80 each).  They are also calibrated for total immersion.  The ASTM 62C has a
range from -38 to +32°C and scale divisions of 0.1.  The ASTM 63C has a range
of -8 to +32 C and also has scale divisions -of 0.1, but they are farther apart
making interpolation more accurate.  If higher temperatures are needed, the
ASTM 64C has a range of 25 to 55°C.

                4.3.1.1.4  Thermocouple systems
                The thermocouple operates on the principle that when two
different metals are joined, a small voltage with a temperature- dependent
magnitude is generated.  By comparing this voltage to the voltage generated by
a second thermocouple in a thermally stable environment of known temperature,
the temperature of the first thermocouple can be determined.  Because of
complex circuitry and problems with conductors, thermocouple systems are no
longer popular transducers for meteorological monitoring.

                Thermocouple pairs are well suited for differential temperature
measurement.  They provide the same voltage for any size wire which makes them
ideal for miniature fast response applications.

       4.3.1.2 Signal Conditioning

       There are a multitude of circuits which will measure resistance.
Usually the transducer and the signal conditioner are purchased as a system,
complete with interconnecting cables.  This is advisable since the range of
resistance vs. temperature is quite large.  Signal conditioning circuits may be
adjusted to conform to individual transducers or transducer pairs.  They may
also be adjusted to a generic or theoretical curve or transfer' function.  It is
important to understand the function of the signal conditioner and to treat it
as a part of a system along with the transducers and the cables.

-------
                                                               Section No. 4.3.1
                                                               Revision No.     0
                                                               Date:   17 Sep  89
                                                               Page:    4  of   4
       4.3.1.3  Aspirated Radiation Shields

       There are  many  kinds of shields, as Figure 4.3.1.1  depicts.  Most of  the
error in measuring  air temperature comes from the shield.   It is also true  that
the magnitude of  the error is largely unknown.   In  Section 4.3.2
SPECIFICATIONS, there  is a detailed discussion about  accuracy of shield
performance.

    The measurement of the temperature of the free  atmosphere at the point
where the shield  intake is located is the goal.  A  shield protects the sensor
fro» radiation  and  provides the mechanical mount for  the sensor on a tower  or
mast.  If the shield is not aspirated, or designed  for effective natural
ventilation, it may become little more than a larger  sensor case resulting  in
the same radiation  errors as an unshielded sensor but with a longer time
constant.  To avoid this difficulty, it is necessary  to draw the air into  the
shield in such  a  way that it is not modified by  the shield temperature but  will
cone to equilibrium with the transducer at some  average time.

       Forced aspiration is the only way to minimize  radiation error for all
conditions.  A  fan  draws air in past the transducer at a speed suitable for
minimum error.  Forced aspiration can be designed to  provide a flow in the
right direction under  all ambient wind conditions.   Insufficient pressure drop
in the fan during strong winds may allow reverse flow to occur transporting the
heat from the fan back to the transducer (Lockhart, 1975).
                                              IT
         A£S Strvtmon Serum
          AES Pmrtllml Pl*m
                                         Telfd^nt Aipirmted
                                          Radiit/on Shi»ld
                                                       Sttvtnton Scremn
                          JL
                         Itrttti Tf>*t/»r Sh^ld
                                   ClimM Motor Aipiruiud Shit Id
                                                       ModHitd Rmmol Shi fid
                         AfS Marine Stottd
                                           fOABOtwfoint
                    A£S DutJ Asoirmrtd Prtftoramf

          Currftt Plmtt Shtmld
Gill AspiruM Shifld
 Figure 4.3.1.1  Examples of various radiation  shields
                 (McKay and McTaggart-Cowan,  1977)

-------
                                                             Section No, 4.3.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  4
4.3.2  SPECIFICATIONS
    The purpose of defining specifications is to give unambiguous meaning to
the terms used by all those who are concerned that the instruments and systems
selected and operated will meet the needs of the application or project.  This
starts with procurement specifications and ends with supporting claims of data
quality.  These specifications provide the basis for receiving inspection and
testing.

    Project and application requirements vary.   To make this handbook as
specific as possible, the examples used will be consistent with those presented
in the On-Site Meteorological Program Guidance for Regulatory Modeling
Applications (EPA, 1987b).  The specifications for temperature are range and
accuracy.   The performance of the radiation shield is not defined by
specification.   There is an implication that the accuracy requirements include
this error source, but if they do there is no way suggested to verify the
performance of the shield.

       4.3.2.1  Delta Temperature (AT)

       The only requirement in EPA (1987a) regarding the vertical temperature
difference is "Errors in measured temperature difference should not exceed
0.003 C/m."  This rate is based on a 0.15 C accuracy for a 50m separation.
The requirement came from a time when AT was traditionally measured between the
lower 10m level and the upper 60m level on a tower.  If a shorter tower is
used, like a 44ra tower, the separation between 10m and 44m, namely 33m, would
show smaller lapse rates and inversions.  If the same accuracy were to be
preserved in measuring the equivalent or representative AT from the shorter
tower, the measurement accuracy had to be better, 0.1 C in this case.  Some
operators went to even shorter towers and to assure an appropriately accurate
measurement system, the requirement was stated as a per meter error.


       The above requirement is impossible to meet with the new 10m towers
including AT.  If the aspirated radiation shields are mounted at 2m and 9m (to
avoid interference with the 10m wind), the requirement is to not exceed an
error of 7 x 0.003 = G.021°C.  This is an accuracy which is hard to prove, let
alone achieve.

       The dynamic range for a AT installation on a 60m tower might be from
-2 C to +15 C.   Convention for positive and negative AT is:

     (-) a lapse rate is the normal decrease of air temperature with
        height limited by the auto convection rate of 3.4°C/100 m or
        0.034 C/ m.  A lapse rate produces a negative AT.

     (+) an inversion is the inverted lapse rate or an increase of air
        temperature with height.  There is no limit for inversion strength.
        An inversion produces a positive AT.

The dynamic range between 2rn and 9m is not much different than that between 10m
and 60m.   During the EPA-BAO experiment in 1982  (Lockhart, 1988),  a pair of AT
sensors was mounted on tower 4 at 2m and 8.6m.   Each sensor was a 10QD Rosemont
platinum RTD in a Young aspirated shield.  On tower 3, also at 2m and 8.6m, a
pair of MRI-YSI linearized thermistors mounted in MRI shields (patterned after
the Young shields) were operated.  Figures 4,3.2.1 and 4.3.2.2 show three days
of 20-minute average AT data and 2m temperature data from tower 3.  Also shown

-------
                                                         Section No.  4.3.2
                                                         Revision No.     0
                                                         Date:   17 Sep 89
                                                         Page:    2  of  4
      -2
         0     6    12   18   0    6    12   13    0    6    12   18   0
                 9/7/82             9/8/82             9/9/82
Figure 4.3.2.1  MRI fiT Data, and AF Difference Data for Iftree Days
      30 -
      25 -•
   O
   9
      20 -
   CL
   E
      15 -•
       10 -
                 9/7/82             9/8/82             9/9/82
Figure 4.3.2.2  Air Temperature at 2 m from the MRI AT Pair

-------
                                                             Section No. 4.3.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  4


is the difference between the two 20-minute average AT measurements from the
two towers.   Note the dynamic range of -1.3°C to +4.5"c (per 6.6m) and the
agreement  between the two different instrument systems.  For the three days the
average difference is -0.02°C with a standard deviation of 0.10°C,  The daily
temperature range was about 17°C.

       The reason the dynamic range is so high for such a small separation
distance Is that the surface is a better radiation receiver and transmitter
than is the air immediately above it.   The surface is almost always hotter or
colder than the air above it.  Convection and mechanical turbulent mixing drive
alaost all of the heat flux between the surface and the air.  The closer the
sensors are to the surface the larger the temperature difference per meter of
separation.   The drawback is that.the closer the AT pair is to the surface the
more sensitive the differential measurement is to local surface conditions or
character.   A black top road will affect a 2-lOrri AT much more than it will
affect a 10-60m AT.  The lower sensor really drives the AT and a 2m temperature
will vary  more than a 10m temperature.

       For these reasons, the suggested procurement specification in the
On-Site guide (EPA, 1987b) in 8.1.3 reads:
     "Range                             -5 to +15 degrees C.
      Relative accuracy (error)         sQ.l degrees C."

       While calibrations and audits of both accuracy and relative accuracy are
usually conducted in controlled environments, the measurement is made in the
atmosphere.   The greatest source of error is usually solar radiation.  Solar
radiation  shield specification is therefore an Important part of the system
specification.  Motor aspirated radiation shields (and possibly naturally
ventilated shields) will satisfy the less critical temperature measurement.  It
is critical that the same motor aspirated shield design be. used for both
sensors used to measure AT.  The expectation is that the errors from radiation
(likely to exceed 0.2 degrees C) will zero out in the differential measurement.
A motor aspirated radiation shield specification might read:
      "Radiation range                    -100 to 1300 W/m2
       Flow rate                          3 m/s or greater
       Radiation error                    <0.2 degree C."

      Data sheets from five manufacturers (listed alphabetically) specify
their aspirated radiation shields as follows:

1.  Climatronics TS-10       Under radiation intensities of 1100 W/m
                             measurement errors due to radiation will not
                             exceed 0.1 C.  Aspiration rate 3 m/s at sensor
                             location.
2.  Met One 076              Radiation error -less than 0.05°F (0.03°C)
                             under maximum solar radiation of
                             1.6 gm-cal/cm2/min (1100 W/m2).
                             Flow rate 500 ft/min (2.5 m/s).

3.  Qualimetrics 815Q-A      Radiation error - 0.05 C during maximum aspiration
                             and full sun.  Air speed 360 ft./min. (1.8 m/s).

-------
                                                             Section No. 4.3.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page;    4  of  4

4.  Teledyne Geotech 327C    Shielding - Under test radiation flux density of
                             1100 W/m  errors caused by radiation are less
                             than 0.1 C.  Aspiration rate 6 m/s at sensor
                             location,

5.  R.  ML Young 43408        Radiation error - under radiation Intensity of
                             1080 W/m2, Ambient temperature - 0.2°C R»S,
                             Delta T - 0.05°C RMS with identical shields
                             equally exposed.  Aspiration rate - 3 m/s.

       It is difficult not to notice the similarity among these very different
designs.   An auditor would need a comparative field test to find the relative
error from solar radiation.  Such a test can be done using a AT sensor pair
Mith two (or more) shields collocated at the same level. * If the transducers
are well calibrated, the relative temperature of the transducers can be known
to 0.02 C.  these transducers in the two aspirated radiation shields will
report the relative performance of the shields to the same relative accuracy.
Hie one which is coolest in  the daytime and warmest at night has the least
radiation error.  Several diurnal cycles with sunny days and clear nights are
required.  Such a test series could Identify the most efficient shield which
could become a standard against which a relative error analysis of any shield
could be made by a CIS method.

       4.3.2.2  Temperature

       The accuracy specification for  temperature is suggested in
EPA (1987b) as

       "Range                            -40 to +60 degrees C.
       Accuracy (error)                 so.5 degress C."

       Some applications such as "PSD" permits without fog problems require an
accuracy of only I degree C.  For locations with winds generally above  1.5 m/s,
a well designed naturally aspirated shield can provide 1 degree C. accuracy.
When the application requires an aspirated radiation shield, the shield
performance requirement should also be included In the specification.   It is
customary to use the lower AT shield to aspirate both the temperature and half
the AT pair.  Some designs develop AT by subtracting one temperature
measurement from the other,  in which case there is only one sensor in the lower
shield.

-------
                                                             Section. No. 4,3,3
                                                             Revision No,    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2
4.3.3  ACCEPTANCE TESTING
    There are two ranges of temperature to consider.  One is the measurement
range and the other is the environmental operating range.  The two might be
similar for remote installations without commercial power or air conditioned
shelters.  The operating temperature range of the signal conditioning circuits
for remote Installations is a function of the radiation shielding of the
electronics and the heat generated by the circuit operation.  The electronics
•ay get both colder and hotter than the air temperature under these
circumstances.  The operating temperature range for a station with an air
conditioned shelter is much narrower than the measurement range.  The receiving
or acceptance test design must consider these factors.  A conventional
temperature chamber is required to control the temperature of the electronics
while the sensors are controlled by a separate thermal environment.

    A test which demonstrates the operation of AT and T sensors connected to
their signal conditioners is recommended.  The test should include at least two
temperatures.  Liquid bath or solid thermal mass devices are recommended to
avoid local gradients in the air.  It is necessary to remove the sensors from
the aspirated radiation shield and place them in a bath while they are still
connected to the signal conditioner.  This may prove difficult to do with some
designs.   Since calibrations and audits are likely to require the same
accessibility of the sensors in the field, it is best to solve this potential
difficulty in the less hostile receiving laboratory space.  Perhaps a statement
of accessibility needs to be included in the specification.

    Experience has demonstrated that thermal stress can cause sensor failure
when baths are used.  The sensors may seem to be sealed, but if a sensor is
submerged in a hot (40°C) bath for enough time to reachoequilibriura and take a
series of measurements, and then submerged in a cold  (0 C) bath, the pressure
change inside the metal sensor cover may draw water Into the element chamber.

    It is prudent to assume that the sensors are not hermetically sealed and
to protect them as much as possible.  Using solid thermal mass devices Is one
way to avoid liquid from wicking or being drawn into the sensor, but it is not
a total protection.  In the above example, room air may be drawn into the
cooling sensor as the pressure inside equalizes with ambient pressure.  The air
drawn in may be saturated at the new cold temperature or may condense some
water vapor.  Whether or not this is a problem depends on how the sensor is
made, but it is best to take what precautions are possible.  If complete
immersion is necessary, wrap the sensors in plastic so no liquid can get to the
interface where the wires come out of the sensor.  Use partial immersion where
possible keeping the interface dry.  Keep temperature changes small and in the
order AMBIENT-COL&4 AMBIENT-»HOT-»AMBIENT.

    Assume the receiving test will use two temperatures, ambient temperature
and an ice slurry.  Assume a water bath will be used for the ambient test.  If
one to three sensors are in the system, wrap them together along with a
thermometer using a rubber band.  Use a Thermos  bottle which has been filled
with water several hours earlier.  The key to accurate temperature measurements
with sensors of different time constants Is in having a thermal mass with
minimal gradients.  The value of a Thermos bottle is its long time constant.
It will tend to keep the temperature of its contents constant, but all it can
do is cause the heating or cooling to be slow.   In time the contents will be at

-------
                                                             Section No, 4.3.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  Z

the same  temperature as the surroundings, if the temperature of the
surroundings Is constant, like an air conditioned room.  When this equilibrium
has been  reached, the water in the bottle will be at the same temperature
everywhere in the bottle and stirring is unnecessary.  Stirring suggests the
need to mix up parts of different temperature.  It is better not  to have parts
of different temperature and this can be achieved by reaching equilibrium with
a well insulated mass.

    Place the sensors and thermometer into the Thermos.  Use a cork or some
cover to  keep ambient air from circulating over the top surface of the water
mass.   After about 30 minutes, assume the sensors and thermometer are in
equilibria™ (they should be, they went in from the same equilibrium
tejperature) an
-------
                                                             Section No. 4,3.4
                                                             Revision No.    0
                                                             Date;   17 Sep 89
                                                             Page:    1  of  1
4,3.4     INSTALLATION
     Each design of aspirated radiation shield has its own installation
requirements.  The manufacturer's manual must be used in addition to the
general guidance given here.

     The installation location is chosen to represent the temperature relevant
to the application.  The height above ground is the first consideration.  If
the temperature is to be used for cliutatologlcal purposes, a 2ai height above a
grass surface will do.  If the temperature is to be used to describe the air
being drawn into a manifold for chemical analysis, the best location is the one
that represents what the manifold inlet "sees."  If a temperature gradient is
to be used to describe the stability of the surface layer, a representative
pair of heights is selected.  Siting is best done with the concurrence of the
person who will be using the data for analytical purposes, the person who will
judge the data to be valid, and the person who will accept the data and the
analysis on behalf of the regulating agency.  Siting by this committee approach
will benefit from any objective knowledge any member might have, but its strong
point is in the mutual understanding of the criteria which were used in making
the selection.

 The second consideration is bias from surrounding structures.  The temperature
that is measured is that of the air which is drawn Into the aspirator.  If
there is a prevailing wind, mount the radiation shield into the wind such that
the wind passes the shield before reaching the tower.  The farther the
aspirator inlet is from the tower, the smaller the angle segment which can
contain the tower heat and the aore mixing with non-heated air by the time the
inlet is reached.  The distance out from the tower should be the maximum
allowed by the mounting hardware.  Special booms for temperature may be
necessary if the design does not provide for siting the inlet at least one
tower diameter from the edge, and if 0.5 C accuracy is expected.

-------
                                                              Section No.  4.3.5
                                                              Revision No.     0
                                                              Date:    17 Sep 89
                                                              Page:     I  of  9
4.3.5  CALIBRATION
    The uanufacturer" s  manual will give  instruction for the adjustment of
signal conditioning  circuits in  response to  some  input  specification.   Usually
what is required  is  a precision  resistor, either  built  in or to be supplied  by
the calibrator  using a  decade box or equivalent.   Such  procedures are  assumed
to have been done and will not be discussed  here.   The  method recommended in
this handbook is  independent of  the manufacturer.   It requires only the generic
transfer  function of the  resistance element  used,  that  is the ohms vs.
temperature  relationship  upon which the  measurement system depends for accurate
performance.

       4.3.5.1  Calibration Equipment and Methods

       The handbook  method requires three stable  thermal mass assemblies with
temperatures known to about 0.1  C.  The  three masses may be one mass used three
times with sufficient time allowed for conditioning to  a new temperature.
Sufficient effort must  go into the determination  of thermal stability  and
accuracy  of  the temperature measuring system used for that determination in
order to  defend the  results of the calibration or audit.   The following will
describe  one solution to  this requirement.   This  solution- is not the only one
but the confidence in stability  and accuracy produced by this solution is an
•example of the  documentation necessary to support claims of accuracy.

    The thermal mass design uses a solid aluminum cylinder,  shown in Figure
4.3.5.1,  chosen for  high  thermal conductivity.  The mass is supported  in the
space inside a  stainless  steel insulated bottle  (see Figure 4.3.5.2) by a
lucite tripod on  the bottom and  by three low thermal conductivity stainless
steel screw  spacers  at  the top.  The lid is  modified to allow transducer cables
to go through.  The  insulating bottle is positioned inside a 2-gallon  insulated
container modified to allow cables to go through  with the top in place as shown
in Figure 4.3.5,3.
Figure 4.3.5.1 Thermal Mass
Figure 4.3.5.2  Insulated bottle

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  9
      Figure 4.3.5.3  Three  mass containers:   Hot,  ambient,  cold (right)

The volumes and masses Involved with this device are listed  in Table 4.3.5.1.
The air space  between the inside of the 2~gallon container  and the outside of
the insulated bottle is filled with either a cylindrical structure with a
heater strip and thermostat "floated" on stainless steel spacers  (#3, hot), air
(#2, aabient) or ice (#1, cold).

            Table 4.3.5.1  -  Details  of  a Solid Mass Thermal  Device
Component

Aluminum cylinder (less holes)
Holes
Air inside quart bottle
Quart bottle
Air outside quart bottle
2-gallon container
Total for tt2 - ambient
. Quart bottle (see above)
Ice-water mixture
Air above ice-water
2-gallon container
Total for #1 - cold
Volume
(cm )
485
58
1,100
931
5,882
10,206
18,662
2,574
4,706
1 , 176
10,206
18,662
Mass
(g)
1,341
0.1
1.3
1,154
7.1
1,600
4,104
2,496.4
4,706
1.4
1,600
8,804
Typical Transducers:
Minco S28F36Y nickel-iron 2.4 7
Rosemont RMT 78-39-7 5 36

-------
                                                           Section No. 4.3.5
                                                           Revision No.    0
                                                           Date:   17 Sep 89
                                                           Page:    3  of  9

    The  ratio of 1,341 grams of thermal mass to about  43 grams of sensors meets
the design goal of a small sensor thermal impact.  The  several layers of
insulation and the minimization of thermal conductivity paths meets the long
time constant goal for the thermal mass.

A test was conducted to document the performance of the thermal mass assemblies
and to show the time required to condition the hot and cold masses and the time
constant of the ambient mass, see Figure 4.3.5.4.   From the beginning of the
test at  a little after 10:00 a.m., when ice was put into il and the heater was
plugged  in for f3, it took about 9 hours for the slope to be flat enough to be
confident that the mass had a homogeneous temperature.  The maximum rate of
change of the ambient mass was 0.01 C/min.  The stability of the measuring
circuit  is shown by the line with triangle symbol.   The thermal conductivity of
almdnum Is 0.5 cal./sec  through a plate one centimeter thick across an area of
one square centimeter when the temperature difference  is 1C.  Steel is 0.1,
rubber is 0.0005 and air  is 0.00005 (Hodgman,  1955).   The aluminum cylinder
exchanges heat with its environment (except for the transducers being tested)
       40
              Thermal Mass  Conditioning Test
                         Meteorological Standards Institute
Figure 4.3.5.4
                12
              Local Daylight Time
Thermal  mass conditioning and response test.
through  450  cm2 of  air, 2.4  cm2 of lucite and 0.1 cm  of steel.   The difference
in temperature between  the holes where the transducers are mounted can be
estiaated  from the  difference  in conductivity within the aluminum and within
the environment outside the  aluminum (99.5% air), or 10,000:1.   If the mean
temperature  of the  cylinder  changes at 0.01 C per minute, the gradient change
between  hole temperatures could be 0.000001°C per minute.

-------
                                                             Section No.  4.3.5
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     4  of  9

    Having created a stable environment, there needs to be an accurate method
of measuring both the relative and absolute temperatures of the thermal masses.
A three transducer data system was designed for this purpose.   The report of
its calibration is shown at the end of this section in 4.3.5.4.

       4.3.5.2  Delta Temperature (AT)

       Calibration of a AT system involves two parts.   One is the matching of
the transducers at zero difference and the other is the gain of the signal
conditioner for a known difference in temperature of element resistance.

       The first part involves placing the AT transducers together in a series
of stable thermal masses.  A stable thermal mass is any mass which is at least
25 tines the Bass of the sensors being conditioned by the thermal mass and
which has thermal gradients of less than 0.01 C/cm throughout the mass.  Start
at cold, soaewhere in the 0°C to 5°C range, and record the system AT output
after stability has been reached in 30 to 60 minutes.   Take readings about five
minutes apart.  When the readings stabilize, average the last five.  Assume the
output reports a difference of -0.02°C.  Then move the two transducers to the
thermal mass at ambient temperature, somewhere in the 15 C to 25 C range.   When
stability has nearly been reached, start taking readings about five minutes
apart.  When the readings have stabilized, average the last five readings..
Assume the output reports a difference of +0.03°C.  Finally, move the two
transducers to the hot thermal mass, somewhere in the 35 C to 40 C range,  and
wait 30 to 6O minutes for stability to be reached.  Take readings about five
minutes apart.  Assuming they are -0.09, -0.07, -0.05, -0.04,  and -0.03,
stability has not been reached.  After another 30 to 60 minutes take another
series of measurements about five minutes apart.  Now they are -0.01, +0.01,
0.00, 0.00 and +0.01.  Stability has been reached and the average of the last
five readings, 0.00, is recorded.


       This test has confirmed that the two sensors are matched to each other
and to the generic transfer function with which the signal conditioning
circuits have been set.  It may be that the matching was done in the circuitry.
It does not matter.  It has been shown that the transducers and their parts of
the circuitry agree with each other at three different temperatures.If
agreement is not within ±0.05 C of the true value of 0.00 C, look to the manual
or the manufacturer for guidance in correcting the problem.  The AT system
should start off with agreement in controlled conditions of much better than
0.1 C if the atmospheric measurements are to approach that accuracy.  The
methods described here for building a stable thermal environment and sampling
the outputs for zero difference are only an example which works.  The only
important criteria is that it be documented in terms of stability, whatever
method is used.


       The second part of the AT calibration sets to tests the gain of ,the
difference amplifier.  Pick a common temperature for the site and substitute a
fixed resistor for one transducer, arbitrarily choose the lower one.  Assume
the transducer is a 100Q platinum type  (see Table 4.3.1.13 and your resister is
108 ±1Q.  Substitute a precision decade box for the upper transducer.  Adjust
the decade box until the AT output is the voltage equivalent of 0.00°C.   If the
range is -5 to +15°C for a 0 to 1 volt output, 0.00 C is 0.250 volts.  The
output now reads 0.250 volts and the decade box reads 107.96Q.  If 107.79Q

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  9


represents 20°C, and if Q.389Q represents a 1°C change (0.03898 for 0.1 °C
change or Q.Q1Q = 0.0257 C), the simulated temperature for both transducers is
(107.96 - 107.79 - 0,17; 17 x 0.0257 = 0.437) 20,437 ±0.026°C or between 20.41
and Z0.46°C.
       If the decade box is changed to 108.35 (107.96 + 0.389), the upper
slaulated transducer Is now 1 C warmer than the lower simulated transducer.
The output should read +1°C or 0.300 volts (1.000 volts * 20 degrees = 0.050
volts/deg.; 0.250 •»• 0.050 = 0.300).  If zero and full scale are to be
challenged, set the decade box to 106.02 (107.96 -  [5x0.389] = 106.02) for a
0.000 voit reading and set the decade box to 113.80  (107.96 +  [15x0.389] -
113.80) for a 1.000 volt reading.  Check the difference in box settings; 113.80
- 106.02 = 7.780; 7.78 + 0.389 » 20.00°C.

       Beware of rounding errors if enough resolution is not carried or
available on the decade box and the output.  Check  the decade box with a good
ohm Meter.  Errors in the tens wheel (when switching from 106 to 113) may be
larger than the smallest wheel.  If the tens wheel  is only good to 1%, the
uncertainty is 0.10Q or 0.26°C.

       4.3.5.3  Temperature

       The temperature calibration may be achieved  concurrently with the AT
calibration.  Each thermal mass should have some accurate means of determining
temperature.  While the actual temperature is .not important for the AT
calibration, it should be recorded on the calibration form.  If the temperature
transducer is not one of the AT pair, it can be placed in a thermal mass at the
same time the AT calibration is being done.

       It nay be that the system does not have a AT measurement but it does
have a temperature measurement.  Remembering that the temperature accuracy
requirement is ±0.5 C, the temperature transducer can be challenged with a much
siapler method.  Liquid baths in a pint or quart insulated bottle with the
transducer and a good ASTM or equivalent thermometer mixing the bath together
will suffice.  Be sure stability has been reached before taking the readings.
Use care or parallax-avoiding devices when reading  the thermometer.

       4.3.5.4  Calibration Report Example

       The following information is reproduced from a report prepared for
Meteorological Standards Institute to document the  accuracy of temperature
instrumentation used on audits.

                4.3.5.4.1  Introduction

                During April 14 to May 8, 1988, a calibration program was
conducted to verify the accuracy of three MINCO 604 ohm RTDs (resistance
thersal devices).  Three RTDs, Minco model S28F36Y  labeled #1, #2 and #3, were
originally calibrated in 1984 and have been in use  for temperature and delta
teaperature auditing during the past four years.  The earlier calibration was a
relative calibration since the only accuracy of consequence to the application
was the intei— relationship of the three RTDs.  The current calibration is both
relative and absolute.

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  9
                4.3.5.4.2  The measurement circuitry

                The three RTDs are connected in series to a battery powered
constant current source of 0.500 mA.  A fixed resistor (668 ohms) of low
thermal sensitivity was also in the series loop as a reference source (REF).
The voltages across each of the four resistors in the current loop were
connected to input channels of an ADC-1 data logger.  A NEC PC-8201A computer
controlled the ADC-1 and collected the data.  A program called ADCT was used to
sample all the channels every 5 seconds and to record the average and standard
deviation for periods of time selected through the program.  Times of 10
minutes, 30 minutes and an hour were used at different times during the
calibration.  The ADC-1 provides an output in tens of millivolts.  If the
voltage across a 668 ohm resistor In a 0.0005 ampere current loop is 0.334
volts or 334 mV, the ADC-1 will output 3340.

                4.3.5.4.3  The conversion of Minco ohms to temperature

                In the Minco Application Aid No. 7, Table 14-604, the nominal
values of resistance for temperatures are given.  Nine sets of temperatures and
resistances in five degree steps from 0 C to 40 C were used to find a
mathematical expression for converting resistance to  C.  A linear regression
was not satisfactory.  The quadratic solution to the regression analysis
predicted the temperature at the nine points with an error of less than 0.01 C.
Table 4.3.5.2 shows the input pairs, the predicted temperature and error, and
the constants found and used.

        Table 4.3.5.2 - Relationship of Minco Resistance to Temperature
y x
	 Minco Table 14-604 	
temperature
C*C5
0.000
5.000
10.000
15.000
20.000
25.000
30.000
35.000
40.000
resistance
(ohms)
604.00
617.98
632.13
646.46
660.97
675.66
690.52
705.57
720.79
A A
y y - y
predicted
temperature
c'o
0.006
4.998
9.994
14.996
20.001
25.005
30.005
35.004
39.994
error
C°C)
0.006
-0.002
-0.006
-0.004
0.001
0.005
0.005
0.004
-0.006
for y » a + bx + ex2 a « -269. 1531523
R = 1.00000009 b = 0.53213288
n = 9 c = -0.00014322177
                A second  step in  the  conversion requires the constant current
to be exactly 0.5 mA  so that  the  recorded voltage can be converted  to
temperature.  The REF resistor is recorded as temperature using  the generic
conversion formula.   The  difference between the recorded value of REF and  the
correct value for 0.5 mA  gives the correction.  To find the correct value  a
series of measurements were made  with the Fluke 8060A In the current loop.  The

-------
                                                           Section No.  4.3.5
                                                           Revision No.    0
                                                           Date:    17 Sep 89
                                                           Page:     7  of  9

least squares straight  line was calculated for the data and both the data and
the best-fit line are shown in Figure 4.3.5.5.  The REF value for 0.5 mA was
calculated and found to be 22.25S deg. C.
   O
   O
   O
    OT
    O>
    
-------
                                                            Section No. 4.3.5
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    8  of  9
                4.3.5.4.5  The test facility

               The three Minco RTDs were taped  to  the  mercury
bulb of the thermometer after all  four devices  were  threaded through a rubber
                          stopper.   The assembly was submerged  in  an ice slurry
                          or in water  in a pint Thermos bottle  (see Figure
                          4,3.5.5).  The data logger signaled when an average
                          was being recorded and the thermometer was read at
                          the same time.   The average  from  the  Minco RTDs
                          represents the middle of the time period while the
                          theraometer  was read  at  the  end of the time period.
                          A linear extrapolation of  the Minco RTD  data to the
                          time of  the  reading of the thermometer provided
                          comparable data.

                          4.3.5.4.6 Results

                               There were 73 thermometer readings  over a nine
                          day period covering a temperature range  of 31.95 F to
                          89.05°F  (~Q.03°C to 31.69  C).  The differences in
                          temperature  between Minco  sensors ttl  and #3,
                          expressed in  C,  and  the thermometer  temperature,
                          expressed in  F,  are  shown in Figure  4.3.5.7.  Sensor
                          f2 was so close to tl that it was not plotted.  The
                          calibration  correction curves for all three sensors
                          are shown on the figure.   The best fit lines from the
                          linear regression analyses are drawn  on  the figure.
                          The coefficients are  listed  in Table  4.3.5.3.

                MINCO  Thermometer  Calibration
                           Meteorological Standards  Institute
Figure 4.3,5.6
 Calibration
      0.5
 O
 
      0.4 -
      0.3 -
      0.2
      0.1
        0
     -0.1 -
                  KEY
              o   |3 (Regression Analysis

              0   $1 (Regression Analysis
            8
            f 3 Correction (-0.265° C)

            §2 Correction (-0.010°C)
     -0.2
            |1 Con-action (-O.OQ2°C)

                                                                       CD
                                                                       •
          30
                               50
                                                   70
                                                                         90
                   ASTM-63F (245453) Temperature  (deg. F)
 Figure 4.3.5.7  Calibration of Hinco sensors (RTDs)

-------
                                                             Section No. 4.3. 5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    9  of  9
              Table  4.3.5.3
Regression output for Minco #1 and #3
versus mercury-in-glass thermometer

Constant (a)
Standard error of y
Coefficient {b)
Standard error of b
Number (n)
Average difference
Standard deviation
Minco #1
0.02262
0.02838
-0.00030705
0.0001 9593
73
0.002
0.028
Minco #3
0.29951
0.02534
-0.00041265
0.00017491
73
0.272
0.025
                The relative calibration, without consideration of an outside
                                                O       9
measurement of temperature, covered a. range of 0 C to 44 C.  There were 254
averages recorded for each of the Minco RTDs.  The average and standard
deviation of the differences between RTDs is shown in Table 4.3.5.4.
                  Table 4,3.5.4  Relative difference analysis
Average difference
Standard deviation
—Minco Sensors (RTDs) 	
tt3 - f 1
0.2634
0.0183
*3 - #2
0.2753
0.0165
f2 - *1
0.0119
                4.3.5.4.7  Conclusion
                Using Minco #1 as a standard of comparison, and adjusting #1
for the bias of 0.002"c as shown in Table 4.3.5.3, the following accuracies of
relative temperature are estimated.
           Minco #1
                 Minco #2
                 Minco #3
  #1  -  0.002  =  temperature  ±0.05oC
  #2  +  0.010  =  temperature  ±0.05°C
  *3  -  0,265  -  temperature  ±0.05°C
Similar tests in 1984 yielded the following corrections:

                            #1  +0.00  = temperature
                            #2+0.03  = temperature
                            #3-0.22  = temperature

showing reasonable stability over four years and a reasonable capability to
duplicate relative calibrations.

-------
                                                             Section No. 4.3.6
                                                             Revision No.    0
                                                             Date:   1? Sep 89
                                                             Page:    1  of  3

4.3,6  OPERATIONS, MAINTENANCE AND QC

       4.3.6.1  Operations


       From the standpoint of quality assurance, routine inspection of the
temperature and AT data will help find problems soon after they occur.  Routine
inspection of the instrument system is also required.  The temperature
equipment is usually free from wear or change, except for the aspirated
radiation shield, which tends to attract homeless critters of all kinds.
Inspection of the temperature shield and transducer should be a part of the
routine site visit and duly noted In the site log book.  If there is no routine
visit,  a weekly or at the very least monthly visit is recommended.  Systems
usually have" some built in calibration feature which substitutes resisters for
the transducers to check or calibrate the signal conditioner.  The site visit
should include a temperature comparison with a simple hand-held thermometer for
reasonableness.  Guard against radiation errors with the hand held thermometer
when working In direct sun light by keeping the thermometer in shade and away
fro» body heat, including hot breath.  The long time constant of
mercury-in-glass thermometers (minutes) may make the inevitable warming from
biasing heat difficult to observe.

       4.3.6.2  Maintenance

       Routine maintenance should Include cleaning the aspirated radiation
shield and verifying its function.  This should not have to be done any more
frequently than the bearing tests of the wind equipment.  Since it is necessary
to get the wind equipment down from the tower for the bearing tests, a cleaning
of the radiation shields at the same time would be prudent and economical.

       4.3.6.3  Quality Control

       The best way to keep a AT system operating "in control" is to usfe a data
quality control inspection for this variable.  The QA Plan should supply the
details of the AT inspection program.

       An important aspect of the inspection is the background of the
inspector.   Ideally, an experienced QC meteorologist should be used.  Lacking
this resource, a training program should be made available to the person who
will routinely perform the data inspections.  The training will point out the
nature of AT data as a function of wind speed, cloud cover, and time of day
(solar angle).

       Additionally, training will point out that a AT value, that is the
difference between two well calibrated and shielded transducers, is just that.
It is not a gradient measurement unless there is reason to believe that the air
between the two transducers is reacting normally to thermal flux.  Cases have
been observed where a 10m to 60m AT averaged in excess of the auto convection
rate for hours.  The easy assumption is that there is an instrument error
because autoconvection rates cannot be exceeded for long periods of time.  The
often unspoken assumption is that the AT transducers are in the same boundary
layer and the difference in temperature represents the stability condition of
the air.  If the site can produce shallow or transitory surface boundary
layers, as can happen with land-water interface regions, one transducer may be

-------
                                                             Section No. 4.3,6
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  3

in one layer and the other transducer in the other layer.  Then the AT value
represents nothing more than single samples in different layers and lapse rate
conclusions are invalid but the data are valid.

       Training will show the normal diurnal cycle from a negative AT (lapse
rate)  In the daytime to a positive AT (inversion) at night (see Figure
4.3.2.1).   An understanding of the physical process will support  the data with
other  observations of weather conditions.  The sun heats the surface much more
than the air above it.  The air at the surface is warmed by the now warmer
surface.   The warmer air expands and rises and mixes with the air it passes.
This unstable convective process continues until the driving force, the surface
waraer than the air above it, Is neutralized.  This can happen by either
changes In the radlatlonal heating of the surface or by the effective cooling
of the surface through the heat removal process described above.  Considering
the strength of radiant heating (sun angle and sky cover) and the strength of
the mixing process (wind speed), the size of the lapse rate (-AT) can be
imagined.

       Conversely at night, the surface is cooled by long-wave radiant  loss to
the cold universe.  As the surface gets colder than the air above it that
surface air is cooled by conduction.  The cooled air is mixed fay  mechanical
turbulence, caused by the wind flowing over surfaces elements, and slowly cools
the air from the surface up.  This very stable process results in the air
above, not yet cooled, being warmer than the air below or an inversion  (+AT),
The size of the inversion results from the amount of surface cooling {sky
cover) and the amount of turbulent mixing (wind speed and surface roughness).
Calm clear nights will have larger inversions than cloudy and windy nights.   If
the wind Is too high, there may not be an inversion at all.  All  the air may be
mixed  so well that there Is no measurable difference between the  two AT
transducers.

       To complicate the picture, temperature measurements also change  as
different air with different conditioning history is blown or advected  past the
measurement site.  Diurnal rhythms can be seen as colder air flows down terrain
features,  such as hills and valleys, at night as a consequence of the kind of
surface cooling described above.  These vertically stable flow fields often
become decoupled from each other.  It Is possible for one transducer to be in
one stream of stable air while the other is In another stream.  When this
happens, the AT value may not represent a temperature gradient as much  as two
separate flow regimes, similar to the boundary layer example.

       The trained inspector learns to see these processes in the AT data and
to recognize physically unusual or unlikely data.  If the data QC inspector
looks  at the data on a weekly schedule, problems will be uncovered shortly
after  they occur, thus avoiding long periods of data loss.  One week Is also
short  enough to allow memory of conditions to be correlated with  the data.
Discrepancy reports originated by the QC Inspector can Initiate the testing and
corrective action. If necessary, by the instrument operator.  If  At
measurements are taken, the purpose is usually the determination  of stability
categories or parameters for use in diffusion models.  This important variable
deserves careful attention and well documented evidence for validity claims.
The accuracy of the measurement Is achievable only with careful calibration.
The accuracy of the data requires as much care given to data inspection between
calibrations.

-------
                                                             Section No.  4.3.6
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  3
       If only temperature is measured, the accuracy requirements are less
difficult and the data QC inspection can be less rigorous.   If data are
collected,  however,  routine inspection is recommended and a, weekly period is
reasonable.   The inspection compares the temperatures to reasonable values for
the week and the nature of the temperature change to realistic patterns.

-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  8


4.3.7  PERFORMANCE AUDIT METHODS

       4.3.7.1  General Considerations

       A performance audit is a measurement made by an Independent method and
person of the accuracy and precision of the performance of the measurement
system.  To make this determination for temperaturemeasurements, knowledge of
the input conditions to the system sensors is required.  It is also necessary
to know what the system will do to these input conditions in producing an
output.  The output is simply the system output.  Given the inputs and the
transfer functions, the output can be predicted.  The difference between the
predicted output and the system output is the error of the system or its
accuracy.  The methodology starts with the ways of controlling and measuring
the sensor inputs or knowing the inputs in an uncontrolled environment.
Prediction of output from the controlled input requires knowledge of the
transfer function but not necessarily its validity.  Knowing the conditions of
the uncontrolled environment does not require knowledge of any Intermediate
steps such as transfer functions.  Temperature in is simply compared to
temperature out.  The method using the latter approach is called the Collocated
Transfer Standard (CIS) method.  The CIS method Is seldom practical for AT.
The best performance audit uses both methods where appropriate.

       A performance audit must follow some written procedures.  Since the
procedures must be relevant to the design of the instrument or system being
audited, only general principles will be described below with some specific
exaaples.  The data froa the audit should essentially fill out an audit form.
It is important, however, for the auditor to be sufficiently experienced to be
able to deviate from the procedure or the form when the pursuit of truth leads
away from the expected.

       The starting point of an audit form is the documentation of the who,
what, where, when, and how the audit values were acquired.

                4.3.7.1.1  Who

                The performance audit report form should contain a space to
Identify the auditor.  The audit report which summarizes the audit findings
should report the names and affiliations of the operators of the system.

                4.3.7.1.2  What

                The form should contain a section to identify the instrument
being audited by manufacturer, model number, and serial number.
Sub-assemblies, such as transducer or RTD and aspirated radiation shield,
should be identified by number.  If they are not numbered, the operator should
be asked to mark them for identification.

                The audit report should contain a list of all the equipment
provided and used by the auditor, including model and serial numbers and time
of last calibration, where relevant.

                4.3.7.1.3  IThere

                The audit form should have a space to show the location of the

-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  8

sensor on a tower, including height.  A sketch is useful to show  the relative
positions of the sensing elements with respect to possible biasing influences,
such as the tower, other sensors and buildings.

                4.3.7.1.4  When

                The date or dates when the audit affected the system operation
should be listed.  The  time when the system or a particular sensor was  taken
"off-line* and put back "on-line" should be listed.  The time, or time  period,
when each data value was taken is vital for the comparison with the system
output.  Implicit in this  is the need for the time the auditor uses to  be
correlated to the time  the operator or the system uses.  The auditor should
rely on the National Bureau of Standards station WWV for correct  time.  Battery
operated receivers, such as the Radio Shack Time Cube, are generally available.

                4.3.7.1.5  How

                The audit  form should either contain a copy of the method used
or reference the method number.  The audit report should contain  copies of the
audit method used.  The methods should be detailed enough to identify each step
in the acquisition of the  audit value and In the conversion of the value to
units compatible with the  system output.  See Figures 4.3.7.1, 4,3.7.2, 4.3.7.4
and 4.3.7.5 for examples of forms and methods.  These figures are intended as
aids for writing specific  methods and drawing the companion form.  These may be
reproduced if they are  relevant.

       4.3.7.2  Temperature Difference  (AT)

                4.3.7.2.1  Sensor control method

                The audit  method should simulate the most complete calibration
method (see 4.3.5).  The first step is  to condition the thermal mass assemblies
to be used to challenge the transducers to a zero difference environment.  If
stirred baths are to be used, be prepared to give ample time for  equilibrium.
The amount of time is not  well predicted by the transducer time constant.
Stability at the 0.01 C resolution  scale is desirable  to back up  claims of
accuracy to 0.1 C."

                The transducers must be removed from the tower along with  their
cables.  At some  installations it will  be difficult to impossible to remove the
cables.  Substitute cables may be used  if care Is taken to make sure the
substitute does not change the output more than 0.018C.  This should be
verified by using a fixed  resistance representing ambient temperature as a
substitute transducer.  Output readings using both the operational cable and
the substitute cable should be recorded on the audit form.

                The AT  audit usually requires much more elapsed tiire than
measurement time.  It  is practical  to have equipment which allows the
transducers to slowly  reach  their equilibrium state and to record this  process
through the entire system. This way the elapsed time  can be used for the  more
labor  intensive audit  variables.

-------
                                                         Section No.  4.3.7
                                                         Revision No,    0
                                                         Date:   17 Sep 89
                                                         Page:    3  of  8
         Temperature Difference - MSI method DLT006 
-------
                 PERFORMANCE AUDIT REPORT by 	

      MEASUREMENT SYSTEM - Temperature difference
           System number	M1	c,__imi_-^^_CCJ_J1._,-J
           Sensor - .                   	ljr^_rmn_-r_n_
           Transducers   ^	L_^_^^^_^,^ro^__,m_^
           Locations       	
                              Section No. 4.3.7
                              Revision No.    0
                              Date:   17 Sep 89
                              Page:    4  of  8

                             	    DLTOO6
           Si gnal_cond x t i oner
           Data channel* 	
      DATE 	/_/	  TIME off line
              on 11na
      TRANSFER  FUNCTION 	
                         Conversion- for taut*-._________.
      TEST RESULTS  •
        DLTOO&-A
      cold    sensor  #      I _______       2^___^_
                •ass     ^     output        output
      time    ohsas   degC   volt   degC   volt   degC
               teat  start
                                 1-2
                                 dif.
                            volt   degC
      average
      ambient   sensor 41
                 mass
      time    ohms    degC
  output
volt   dagC
  output
volt   degC
    dif.
volt   degC
      avefage

      hot        sensor #
                 mass
      time    ohms    degC
  output
volt   deaC
 2	
  output
volt   deaC
    1-2
    dif.
volt   deaC
      aver-age                •     -        '  '        '            	

        DLTOOe-B
                  Tl        T2A       Tl       T2      S    B-A
      sensor # 1	2	 1-2    1	2	   1-2
                mass       mass   dif*   output    output   dif.   dif.
      time  ohms degC ohms degC degC  volt degC volt  degC degC  dsgC
      aver-age                   -      •                           ____

                  T2       T1A       T2       Tl      B    B-A
      sensor  # 1	,	    2	 1-2    1	2	   1-2
                mass       mass   dif.   output    output   dif.   dif.
      time  ohms  degC ohms degC degC  volt degC volt  degC degC  degC
      aver-aae
Figure 4.3.7.2  Audit  form for the temperature difference method

-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  8


                The transducers are challenged with a known zero difference at
three temperatures.  This shows how well the transducers are matched and how
well they follow the generic transfer function.  The acquisition of these three
data points may take four hours.  Some judgment is required to identify the
point at which stability is achieved.  One clue is the reversal of a
progression of data points.  If, for example, the five minute average AT values
are 0.09, O.07, 0.06, 0.06, 0.05, 0.05, 0.06, 0.05, the last 0.06 can be taken
as a signal that equilibrium has been reached.  The reported value could be the
average of'five stable points (0.05) or the final stable value.

                After the zero difference test has been completed, the gain of
the difference circuit can be challenged by using the substitute resistance
method.  This method is described in 4.3.5.  It is possible to use thermal
•asses at different temperatures, but it is not recommended because of the
uncertainties of stability and four transducers measuring two temperatures to
an accuracy of at least 0.025 C.  Once the matching at zero difference has been
established, the gain is adequately verified by normal electronic circuit
procedures.

                It is not practical to mount a second pair of sensors and
aspirated radiation shields for collocated testing.  The interference problems
with the aspirators are hard to overcome.  The physical problems associated
with mounting parallel instruments are large compared to the value of the
method.  Considerable, but much less, effort is required for the Sensor Control
method which provides numbers with acceptable confidence in their accuracy.

       4.3.T. 3  Temperature

                4.3.7.3.1  Sensor control method

                Usually the temperature transducer, if it is different from one
of the AT pair, can be included in the thermal mass with the AT pair, or in
another thermal mass at the same time as the AT pair is being tested.  Timing
for stability can include temperature with AT as though it were the same test.
The big difference, however, is that the temperature transducer output is
coapared to the thermal mass transducer output as the audit value.  A
calibration of the thermal mass transducer is the key to the claim for accuracy
given the challenged system.

                If there is no AT system being audited,  a simpler method is
appropriate for the temperature system.  A two point check using an ice slurry
and an ambient bath is acceptable.   Two insulated bottles (pint or quart size)
with a cork to support a calibrated thermometer or a calibrated electric
thermometer are required (see Fig.  4.3.5.6).Stability is easier to find since
the readings are only taken to the nearest 0.1 C.

       4.7.3.2  GTS Method

       It is both practical and recommended to use the CTS method for
teaperature audits.  The temperature transducer and its aspirated radiation
shield (or even naturally aspirated shield) is usually located at an easily
reached elevation.   A CTS such as the Assmann Psychrometer shown in Fig.
4.3.7.3 can be located near the temperature sensor.   It should be exposed so
the wind reaches the CTS without bias error from other structures.  If the wind

-------
                                                             Section No, 4.3.7
                                                             Revision No,    O
                                                             Date:   17 Sep 89
                                                             Page:    6  of  8
during the audit is passing through a tower to reach the temperature sensor,
the CTS should not be exposed to sample the same biased air temperature.  The
presumption is that the temperature sensor represents the air temperature.  Any
error from siting is a part of the measurement error.  The CTS should be
mounted to avoid all bias, if possible.

                The CTS method should be
used as an additional challenge to the                                  ,
temperature system.  The two point Sensor
Control method is a challenge to the
transducer and signal conditioning circuit.
The CTS method is a challenge to the
radiation shield at one point on the range
scale.  A one point challenge of a
temperature system   provides no
information about other temperatures.  It
may be that the operator calibrated the
system at one point, perhaps the same
ambient temperature as exists during the
audit.  There could be a slope error which
causes large errors"at near freezing.
Having an accurate temperature measurement
near freezing, accurate to 0,5 C, can be
valuable as it relates to other sensors,
such as wind.   If the wind vane does not
show any direction variation and the
temperature system reports 0.0 C, there is
a good chance that ice is on the direction
vane bearing assembly.  If the temperature
                o           o
system reports 3 C, with a 3 C error
because only one point was used in
calibration and audit, a different estimate
of the direction vane problem is necessary.
If audit or calibration records exist
showing the full range performance of the
temperature circuit, a one point spot check          Figure 4,3.7.3
with a CTS is useful.                              Assmann psychrometer

-------
                                                         Section No. 4.3,7
                                                         Revision No.    0
                                                         Date:   17 Sep .89
                                                         Page:    7  of  8
             Temperature  - MSI method TEH005  (version 8/1/84)

       This method provides for a comparison of a temperature transducer
       with electric output to a calibrated transducer in a slowly
       changing thermal  mass at three different temperatures*  The audit
       equipment  includes three thermal mass assemblies.  Each is a
       cylinder of aluminum (6.4 cm diameter by 17 cm long, 1,341 g in
       4B5  cc) with holes for different kinds of sensors.  This 485 cc
       mass is suspended in the l,lOO.cc inside volume of a stainless
       steel  Thermos bottle, which is inside the 8500 cc inside volume
       of an 18,500 cc cylinder 
-------
                                                           Section No.  4.3.7
                                                           Revision No.    0
                                                           Date:    17 Sep  89
                                                           Page:     S  of  8
                PERFORMANCE AUDIT REPORT by	      TEMOO3

     MEASUREMENT SYSTEM - Temperature
          System number ._.„__	^	_-^__,__.- , -   --^--^-.-.-.-.-^.^^
          Sensor    ,		r^::c:_::::t:::c = ,-.mjJ^,::c:::_ a::i:=im=,:,,-fT	,„„	
          Tr ansdue er    _^^___,___	___•___«___„___
          Location      		                    	 	
          Signal  conditioner
          Data channel  	
     SATE 	/	/	 TIME o-f-f line	on line	test start  _.
     TRANSFER FUNCTION 
                        Conversion -formula
     TEST RESULTS

       TEMQO5A         cold                        ambient
                mass   output   dif.                mass   output    di-f.
     time ohms degC'volt degC  degC      time ohms degC volt degC   degC
          average              „„,„.,..

                        hot
                mass   output   dif.
     time ohms degC volt degC  degC
          averaae
       TEMOO5B
          time           Assroann          output          dif-ference
                       dry   wet  RH
                      degC  degC   %     vole   degC           degC
Figure 4,3.7.5  Audit Form for the Temperature Method

-------
                                                             Section No,  4.3.8
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1
4.3.8  ESTIMATING ACCURACY AND PRECISION
    Section 4.1.5 contains a detailed discussion of methods of estimating
accuracy,  precision and bias using wind speed and wind direction as examples.

    Temperature transducers [as differentiated from temperature sensors which
•ay. include the radiation shielding) may be calibrated or audited by exposing
them to a controlled environment such as a wet or dry thermal mass.  The
ta^jerature of the mass is known either by an installed calibrated transducer
or by collocating a calibrated transducer in the mass.  The collocation
alternative requires assurance that the mass is at a- homogeneous temperature.
The installed option requires assurance that the installed location is
representative of the homogeneous mass temperature.  When such a method is
used, traditional statistical or metrological methods may be used to estimate
accuracy.   The precision of the method is the standard deviation about the
series mean value of repeated Measurements in the constant and controlled
environment.  Such methods are capable of achieving accuracies of 0.1 degree C
and precision of 0.05 degree C or less.

    When a collocated transfer standard (CTS) method is used, considerable care
is required in stating the accuracy of the calibration or audit.  The different
exposures in the atmosphere of different transducers with different time
constants in different radiation shields puts a larger uncertainty on the
cQ«par4son than is found with a controlled environment.  If, for example, a 2
aeter temperature instrument is compared to an Assmarm thermometer mounted
nearby at the same height, the accuracy claim might be no better than the sum
of the two different instrument accuracies.  The accuracy of the method might
be 1 degree C if each instrument is capable of a measurement accuracy of 0.5
degree C.   There is reason to consider such a method as a comparative
measurement rather than an audit or calibration.

    It is possible that a CTS method can have greater accuracy.  What is needed
is a body of data which sets the functional precision of the CTS method by
finding the best one can expect from collocated temperature Instruments.  A
co»canion requirement is a body of data which compares different radiation
shields as a function of radiation intensity, wind speed and wind direction
relative to the orientation of the aspirator motor.

-------
                                                             Section No. 4,3.9
                                                             Revision No,    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1
4.3.9  BEFEREKCES
Brock,  F. V. and C. E. Nicolaidis, 1984:  Instructor's Handbook on
    Meteorological  Instrumentation,  NCAR/TN-237+1A, National Center for
    Atmospheric Research,  Boulder, CO.

Cole, H. L. . 1978:  Air  temperature and differential temperature measurement
    using 1C temperature sensors.  Preprints of Fourth Symposium on Met.
    Observ. and Inst., Amer. Meteor. Soc., Boston, MA., pp. 25-30.

EPA, 1987a:  Ambient Monitoring Guidelines for Prevention of Significant
    Deterioration  (PSD),  EPA-450/4-87-007.  Office of Air Quality Planning and
    Standards, Research  Triangle Park, NC.

EPA, 1987b:  On-Site Meteorological Program Guidance for Regulatory Modeling
    Applications, EPA-450/4-87-013.  Office of Air Quality Planning and
    Standards, Research  Triangle Park, NC.

Hodgman, C. D. , editor in chief, 1955:  Handbook of chemistry and physics.
    Chemical Rubber Publishing Company, Cleveland, OH.

Lockhart, T. J., 1975:   Variable errors in operational data networks.
    Proceedings of  the 3rd Symposium on Met. Observ. and Inst., Amer. Meteor.
    Sac., Boston, MA. , pp. 91-96.

Lockhart, T. J. and M. T,  Gannon,  1978:  Accuracy and precision of field
    calibration of  temperature difference systems.  Proceedings of the National
    Conf. on Qual.  Assur.  of Environ. Meas., Denver, CO., Nov. 27-29.

Mason,  C. J. and H, Moses, 1984:   Atmospheric Science and Power Production,
   Darryl Randerson, Ed.,  DOE/TIC-27601, pp. 103-109

McKay,  D. J. and J. D. McTaggert-Cowan, 1977:  An intercomparison of radiation
    shields for auto stations.  World Meteorological Organization Publication
    No. 48O, pp. 208-213.

-------
                                                             Section No. 4.4.0
                                                             Revision No.    0
                                                             Date;   17 Sep 89
                                                             Page     1  of  7

                QUALITY ASSURANCE FOR PRECIPITATION MEASUREMENTS

4.4.0  DiTHODOCTION

    By definition, "The total amount of precipitation which reaches the ground
in a stated period is expressed as the depth to which it would cover a
horizontal projection of the earth* s surface if there were no loss by
evaporation or run-off and if any part of the precipitation falling as snow or
ice were melted"  (WHO, 1971).  In any method of precipitation measurement, the
ai» should be to obtain a sample that is representative of the fall in the
area.   At the outset, it should be recognized that the extrapolation of
precipitation amounts from a single location to represent an entire region is
an assumption that is statistically questionable.   A network of stations with a
density suitable to the investigation is preferable.

4.4.1  TYPES OF INSTBUlfENTS

    Precipitation collectors are of two basic types:   nonrecording and
record ing!

       4.4.1.1 Nonrecording Gages

       In its simplest form, a precipitation gage consists of a cylinder, such
as a can «ith straight sides, closed at one end and open at the other.  The
depth of the liquid in the can can be measured with a measuring stick
calibrated in subdivisions of centimeters or inches (Figure 4.4.1).

    To obtain greater resolution, as in the case of the standard 8-lnch gage
made to NWS Specification No. 450.2301, the gage is constructed with a ratio of
10:1 between the area of the outside collector cylinder and the inside measuring
tube.   The funnel attached to the collector both directs the precipitation Into
the tube and minimizes evaporation loss.  Amounts in excess of two inches of
rainfall overflow Into the outer can, and all measurements of liquid and melted
precipitation are made in the measuring tube with a measuring stick.

    The automatic wet/dry precipitation collector, available in several
designs, represents a specialized nonrecording instrument designed for programs
involving the chemical and/or radioactive analysis of precipitation.  The
collector is built with a sensor that detects the onset and cessation of
precipitation and automatically releases a lid to open and cover the collector.
In one design, the lid can be made to remain open during either wet or dry
periods.  Another model is made with two collectors;  the lid is made to cover
one bucket during periods of rain and snow (Figure 4.4.2).   In equipment of
this kind involving precipitation chemistry,  the volume of water in proportion
to the constituents collected with the water is important,  so evaporation must
be kept to a minimum {see EPA, 1985).

       4.4.1.2 Recording Gages

Recording gages are of two basic designs based on their operating principles:
the weighing-type gage and the tipping bucket-type gage (Figure 4.4.3).   The
former,  wben made to NWS Specification No. 450.2201,  is known as the Universal
gage,  indicating usage for both liquid and frozen precipitation.  There are
options for the remote transmission of signals from this type of gage.  The

-------
                                                              Section No.  4.4.0
                                                              Revision No.     0
                                                              Date;    17 Sep 89
                                                              Page     2   of  7
standard Mational Weather Service Tipping Bucket Rain Gage  is  designed  with a
12-inch collector funnel that directs the precipitation  to  a small  outlet
directly over two equal compartments, or buckets, which  tilt in sequence with
each 0.01 Inches of rainfall.  The motion of the buckets causes a mercury
switch closure.  Normally operated on 6 V d.c., the contact closure can be
•onitored on a visual counter and/or one of several recorders.   The
digital-type Impulse can also be used with computer-compatible equipment.
Figure 4.4.1
A typical non-recording
Rain Gage (Belfort
Instrument Co. )
Figure 4.4.2 Automatic
wet/dry precipitation
collector.
 Buctet
 Platform
                                         Tipping
                                         Bucket
Figure 4.4.3
A Typical Weighing Rain Gage (left) and Typical  Tipping
Bucket Rain Gage (Belfort Instrument Company).

-------
                                                             Section No. 4.4,0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     3  of  7
       4.4.1,3  Instrument Characteristics
       The «ost accurate precipitation gage is the indicating-type gage.
However,  the record ing-type gage measures the time of beginning and ending of
rainfall  and rate of fall.  The Universal weighing gage incorporates a chart
drum that is made to rotate by either an 8-day spring-wound clock or a
battery-powered clock.   Recent developments include a unit with a quartz
crystal mechanism with gear shafts for a wide range of rotation periods from
half a day to one month.

       The weighing gage is sometimes identified by the name of its designer
(Fergusson) and coaes with one of two recording mechanisms.  In the single
traverse  unit, the pen aoves from the base of the drum to the top, typically a
water equivalent of 6 inches.  In a dual traverse unit, the pen moves up and
then down for a total of 12 Inches of precipitation.  A variation of the
weighing  gage, a "high capacity" design with dual traverse, will collect as
much as 760 nun or 30 Inches.  To minimize the oscillations incurred by strong
wisds on  the balance mechanism, weighing gages are fitted with a damper
iimersed  in silicone fluid.  By incorporating a potentiometer in the mechanism,
the gage  is capable of providing a resistance or, as another refinement, a
voltage proportional to the amount of precipitation collected.  Linearity of
response  is usually a factory adjustment involving the use of calibrated
weights to simulate precipitation amounts.  In spite of manufacturer's
specifications, it is doubtful that the gage can resolve 0.01 inches,
especially when the bucket is nearly empty.

       In the tipping bucket gage, the balance of the buckets and the leveling
of the bucket frame are critical.  Low voltage at the gage is imperative for
reasons of safety.  Power is typically 6 V d.c..  The signal is provided by a
switch closure each time the bucket assembly tips (0.01 Inches of rainfall per
bucket).   Rain rates are calculated from an event recorder with pens energized
sequentially to improve resolution.  The tipping bucket (a mechanical device)
takes time to tilt from one position to the next.  When the rate of fall is
high, there is spillage and the unmeasured precipitation falls into the
reservoir.  Where there is a need for greater accuracy, the collected water is
measured  manually, and excess amounts are allocated proportionately in the
record.  The accuracy of the gage is given as 1 percent for rainfall rates of 1
in/hr or  less; 4 percent for rates of 3 in/hr; and 6 percent for rates up to 6
in/hr.

       4.4.1.4 Accessories - Windshields and Heaters

       Accuracy of measurement for all types of gages Is influenced perhaps
more by exposure than by variations in design.  Windshields represent an
essential accessory to improve the catch of precipitation, especially snow in
windy conditions.  The improved Alter design, made of 32 free-swinging but
separated leaves supported 1/2 inch above the level of the gage collecting
orifice,  is an effective way to improve the catch.  In a comparison of shielded
and unshielded 8-inch gages, it has been shown that at a wind speed of 5 mph,
the efficiency of the unshielded gage decreases by 25 percent, and at 10 mph,
the efficiency of the gage decreases by 40 percent  (Weiss, 1961).

       In below freezing conditions when the catch  in a gage is snow or some
other form of solid precipitation, it is necessary  to remove the

-------
                                                             Section No.  4.4.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page     4  of  7

collector/funnel of nonrecording gages and the funnel in recording gages.   Some
instruments are available with built-in heater elements that are
thermostatically controlled.  An effective heater for conditions that are not
too severe is an incandescent lamp installed in the housing of the gage.
Caution should be exercised, however, as too great a heat will result In
evaporative loss.

4.4.2  SPECIFICATIONS

       4.4.2.1 Precipitation Data Requirements'

       In research studies, especially those related to acid rain,  the
instrument used most frequently is the Automatic Precipitation Collector with
one or two collecting buckets and a cover to prevent evaporation.  In
operational activities, the choice is between the weighing gage and the tipping
bucket gage.  For climatological surveys, the choice might include one of the
above gages as well as a nonrecording type gage.  The use of a windshield is
reco«mended to minimize the errors that result from windy conditions if the
application requires maximum accuracy.

       The precipitation measurement made in air quality monitoring stations is
frequently used for descriptive purposes or for episodal analysis.   If the
effort required to achieve the level of accuracy specified by most
manufacturers of electrical recording gages is more than the application of the
data can justify, a tolerance of 10 percent may be adequate.

       4.4.2.2 Procurement

       In purchasing a suitable precipitation measuring system, specify the
type that fits the data application and include a requirement for accuracy
consistent with that application.  A variety of gages are available
eo»«ercially.  In general,  the standards established by NWS specifications
result in the fewest problems.  For example, there are numerous 8-inch gages
available, but those following NWS specifications are made only of brass and
copper, are aore durable, and are reported to rupture less frequently under
extended freezing conditions than those made of galvanized steel.

       The procurement of a weighing type gage should include a tripod mounting
base as well as a set of calibration weights.  For locations that may not be
readily accessible, or for  locations with heavy precipitation, the bucket of
the weighing gage should have an overflow tube.   Refer to Section 4.4.3.2 for
antifreeze specifications.  If the resolution of time is not too important,
recording rain gages of the drum type can be obtained with monthly rather than
weekly mechanisms.  Unless  the tipping bucket gage is equipped with a heater,
it is of no use for frozen precipitation.

       4.4.2-3  Acceptance Testing

       Except for visual inspection, nonrecording gages do not require
acceptance testing.  The weighing gages should be assembled and given a quick
"bench-top" calibration check using standard weights or a measured volume of
water.  In addition, the clock mechanism supplied with the gage should be
checked for at least a couple of days, preferably a week.  The tipping bucket

-------
                                                             Section Ho. 4.4,0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     5  of  7

gage should also be bench tested, primarily to be certain that the bucket
mechanisn assembly is balanced and that the switch is operational.

       4.4.2.4 Calibration

       Bench calibrations should follow the recommendation of the manufacturer.
The electrical output gage or the drum recording gage measures weight, whether
total weigjht in the case of the "weighing gage" or increments of weight in the
case of the tipping bucket gage.  Density of water is assumed so the weight can
be expressed in units of volume or depth assuming the area of the collector
opening.   Calibrations of the measurement apparatus can be based upon the
introduction of known volumes of water.  The area of the collection surface
must be known for the volume collected to be expressed as a depth.  For
example,  an "eight inch" collector may feed a tipping bucket which tips when
7.95 cc of water has arrived.  If this volume of water is to represent 0.01" of
rainfall,  the effective collection area must be 48,51 square inches, from the
following calculations:

                  7.95 cc - 0.485 In.3  - 0.01  in. » 48.51  in.2

If the area is a circle, the diameter should be 7.86 inches.
                         (48.51/Jt)1/2 » 3.93 in. radius

For rate-sensitive systems such as the tipping bucket, the rate of simulated
precipitation should be kept less than one inch per hour.  Calibrations require
properly leveled weighing systems (gages) whether on the bench or in the field.

4.4.3     OPERATIONS

       4.4.3.1 Installation

       Refer to Section 4.0.4.4.2.4 which provides some siting guidance for
precipitation measurement.  The support,  or base,  of any gage must be firmly
anchored,  preferably on a level surface so that the sides of the gage are
vertical and the collector is horizontal.  The collector can be checked with a
carpenter's level placed at two intersecting positions.  The level of the
bucket assembly on the tipping bucket gage is also critical and should also be
checked along its length and width.

       Once the weighing gage is installed, the silicone fluid should be poured
into the damping cylinder as required.   The pen of the drum recording type is
inked to less than capacity because the ink used is hygroscopic and expands
with increasing humidity, easily spilling over the chart.  The final
calibration check with standard weights or suitable substitute should be made
at this ti»e.  To check the operation of "the tipping bucket, the best approach
is to put a. known quantity of water in a can with a small hole so that the slow
flow can be timed.  It may be necessary to adjust the set screws, which act as
limits to the travel of the tilting buckets.   The average of a minimum of ten
tips should be used.  Adjustment is required if a 10 percent or greater error
is found or if greater accuracy is needed.

       4.4.3.2  Field Operation of a Precipitation Measurement System

       Calibration checks for weighing and tipping bucket gages using the
techniques described above are recommended at 6-month intervals.   Nonrecording
gages, whether alone or in a network,  should be read daily at a standard time.

-------
                                                             Section No, 4.4.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     6  of  7


       Although the weighing gage is used for liquid and frozen precipitation,
it requires some special attention for winter operations.  First, the funnel
must be removed when snow is expected.  Second,  the bucket must be charged with
an antifreeze, 24 oz of ethylene glycol mixed with 8 oz of oil.  The weight of
this mixture represents the baseline from which precipitation amounts are to be
noted.   The bucket should be emptied and recharged when necessary, at about 5
inches in the Universal gage, and at about 10 Inches in the punched tape gage.
If the antifreeze mixture is classified as hazardous or environmentally
sensitive* care must be taken in the disposal process.   All operational
activities should be recorded in the station log.

       4.4.3.3 Preventive Maintenance

       Possible leaks in the measuring tube or the overflow container of the
gage are easily checked.  The receptacles are partially filled with water
colored with red ink and placed over a piece of newspaper.  This procedure is
especially applicable to the clear plastic 4-lnch gage which is more easily
damaged.  Repairs are performed by soldering the 8-lnch gage and by applying a
solvent to the plastic.

       A number of pens, some with greater capacity than others, can be used
with the Universal gage.  All require occasional cleaning, including a good
soaking and wiping in a mixture of water and detergent.  After Inking problems,
the next source of trouble is the chart drive; but these problems can sometimes
be avoided by having the clock drive lubricated for the environmental
conditions expected.  It is a good practice to have spare clocks in stock.

       Routine visual checks of the performance of weighing type gages should
be made every time there is a chart change.  The time and date of change, and
site location should be documented.  Routine maintenance should Include  inking
the pen and winding the clock.  Battery-powered chart drives will require
periodic replacement of batteries based on either experience or manufacturer* s
recommendations.  All preventive maintenance activities should be noted  in the
log book.

4.4.4  PERFORMANCE AUDIT METHODS

    Audits on precipitation measuring systems need be no more frequent than
every 6 months.  The irregular occurrence of precipitation makes the use of a
CIS impractical.  The performance audit should depend upon the challenging of
the gage with amounts of water known to an accuracy of at least 1 percent of
the total to be used.  This method will provide an accuracy of the measurement
system but not the collection efficiency of the gage in natural precipitation.
For tipping bucket gages use a rate of less than one inch per hour and an
amount which will cause a minimum of ten tips.

    For weighing gages, it is more convenient to use calibration weights to
challenge the weighing mechanism rather than using the gallons of water
necessary for full scale testing.

    All types of precipitation gages should be measured to determine the
effective collection area.  This measurement is only required once but the
difficulty of measuring the area of a slightly out-of-round collector may
require several samples to accurately find the area.

-------
                                                             Section No. 4.4.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     7  of  7
4.4. S     REFERENCES
EPA, 1985:  Quality Assurance Handbook for Air Pollution Measurement Systems,
    Vol.  V, Precip. Measurement Systems, EPA-6QO/4~S2/042a.  Office of Research
    and Development, Res. Triangle Park, NC 27711.

NCAR, 1984:  Instructor's Handbook on Meteorological Instrumentation, F. V.
    Brock, Editor. NCAR Technical Note, NCAR/TN-237+1A.

Weiss,L . L. , 1961:  Relative catches of snow In shielded and unshielded gages
    at different wind speeds. Monthly Weather Review, Vol. 89.

MMD, 1971:  Guide  to meteorological instrument and observing practices. World
    Meteorological Organization No. 8TP3, 4th edition, Geneva, Switzerland.

-------
                                                             Section No, 4,5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    1  of 10
       QUALITY ASSURANCE FOR RELATIVE HUMIDITY OR DEW POINT TEMPERATURE

4.5.0  INTRODUCTION

       Humidity Is a general term for  the water-vapor content of air.  Other,
•ore specific, terms for humidity Include:  absolute humidity, relative humid-
ity, specific humidity, mixing ratio,  and dew point (Huschke, 1959).  This
section discusses the measurement of relative humidity and dew point.  Relative
humidity (RH) is a dlaensionless ratio of the actual vapor pressure of air to
the saturation vapor pressure at a given dry bulb temperature.  Dew point Is
the temperature to which air must be cooled, at constant pressure and constant
water vapor content, to be saturated with respect to liquid water.  Frost point
is the temperature below 0°C at which  air is saturated with respect to Ice.
4.5,1
TYPES OF INSTRUMENTS
       There are many ways to measure the water vapor content of the atmo-
sphere.  These can be classified in terms of the six physical principles
(Middleton'and Spllhaus, 1953) listed in Table 4.5.1.  Examples of instruments
for each technique are provided.
                Table 4.5.1  Principles of Humidity Measurement
                 Principle
          Reduction of temperature
            by evaporation
          Dimensional changes due to
            absorption of moisture,
            based on hygroscopic
            properties of materials
          Chemical or electrical
            changes due to absorption
            or adsorption
          Formation of dew or frost
            by artificial cooling

          Diffusion of moisture
            through porous membranes
          Absorption spectra of water
            vapor
                                     Instrument/Method
                                  psychrometer
                                  hygrometers with sensors
                                    of hair, wood, natural
                                    and synthetic fibers
                                  electric hygrometers
                                    such as Dunmore Cell;
                                    lithium, carbon, and
                                    aluminum oxide strips;
                                    capacitance film
                                  cooled mirror surfaces
                                  diffusion hygrometers
                                  Infrared and UV absorp-
                                    tion; Lyman-alpha
                                    radiation hygrometers
       Instruments such as diffusion hygrometers that involve the diffusion of
moisture through porous membranes are used primarily In research programs.  The
same is true of instruments that utilize the absorption spectra of water vapor,
such as infrared and ultraviolet hygrometers, and Lyman-alpha radiation hygro-

-------
                                                             Section No.  4,5,0
                                                             Revision No.     0
                                                             Date:    6 Aug 89
                                                             Page:    2  of 10
meters.   This class of Instrument requires frequent attention and represents a
major investment in procurement and maintenance costs.

      Psychrometry identifies a basic technique for
deriving both relative humidity and dew point temperature
from a pair of thermometers—a dry bulb thermometer that
measures the ambient temperature, and a wet bulb thermo-
meter.   The reservoir of the wet bulb thermometer is
covered  with a muslin wick.  When the wick is moistened
and the  thermometer ventilated, the indicated temperature
is related to the amount of evaporative cooling that takes
place at the existing ambient temperature, water vapor
partial  pressure, and atmospheric pressure.

      The temperature sensors in a sling psychroraeter
(Figure  4.5.1) are usually mercury- or alcohol-filled
thermometers.  The same is true of portable motor-
operated psychroiheters (Figure 4.3.7.3), but the psychro-
metric principle has been used with sensors made of
thermocouples, wire-wound resistance thermometers, therm-
istors,  and bimetal thermometers.  Relative humidity and
dew point are easily determined by observ- ing the
difference between the dry bulb and the wet bulb—the wet
bulb depression—and then referring to psychrometric
tables,  charts, or calculators.  One must be certain to
use computed values for the atmospheric pressure range of
the location where the observation is taken.
                                                               Figure 4.5.1
                                                                  Sling
                                                               psychrometer
       More Measurements of atmospheric water vapor have
probably been made with the sling psychrometer than by any
other manual method.   When properly used and read, the
technique can be reasonably accurate, but it is easily misused.  The most
important errors are from radiation, changes during reading, and parallax.  The
Assaann psychrometer continuously aspirates the thermometers and protects them
                                      from radiation which allows time and
                                      accessibility for a careful reading to
                                      avoid parallax (a parallax avoiding guide
                                      to keep the eye perpendicular to the
                                      meniscus is best, see Figure 4.5.2).  For
                                      good accuracy, particularly where a
                                      variety of observers are taking measure-
                                      ments, an Assmann or equivalent type
                                      psychrometer is recommended.  One should
                                      use the psychrometric tables with dew
                                      point values for the altitude (pressure)
                                      where measurements are being made.

                                           Hygrographs, which record relative
                                      humidity, or hygrothermographs,  which
                                      record both relative humidity and temper-
                                      ature, usually incorporate human hair as
  Figure 4.5.2  Assmann psychro-    '  the moisture-absorbing sensor.  Other
   «rt«r with parallax guides,        instruments with sensors that respond to
                                      water vapor by exhibiting dimensional

-------
                                                             Section No.  4.5,0
                                                             Revision No.     0
                                                             Date:     6 Aug 89
                                                             Page:     3  of 10
changes are available.  They are made with elements such as wood, goldbeater's
skin (an animal membrane}, and synthetic materials,  especially nylon.

       Instruments that utilize the hygroscopic characteristics of human hair
are used »ost frequently, primarily because of availability.   The hygrograph
provides a direct measure of relative humidity In a portable instrument that is
uncomplicated and Is relatively inexpensive.   There are limitations in accuracy
below 20 percent relative humidity and above 80 percent that may be unaccept-
able,  as well as limitations for applications at low temperatures.  Atmospheric
Environment Services of Canada has found that Pernix,  a specially treated and
flattened hair element, can be used at temperatures below freezing without
serious errors.  The hygrothermograph made to an NWS specification incorporates
huaan hair as the humidity sensor and bourdon tube (a curved capsule filled
with alcohol) as the temperature sensor.

       Dew point hygrometers with continuous electrical outputs are in common
use for Monitoring.  One dew point hygrometer was originally developed for air
conditioning control applications under the trade name Dewcel (Hickes, 1947)
and was adopted to meteorological use CConover, 1950).  From the trade name,
the generic term dew cell has evolved that now identifies an instrument made by
several manufacturers.  This device determines moisture based on the principle
that for every water vapor pressure there is an equilibrium temperature at
w&ich the saturated salt solution neither absorbs nor gives up moisture to the
surrounding atmosphere.

       The dew cell, also known by the trade name Dew
Probe,  consists of bifilar wire electrodes wrapped
around a woven glass cloth sleeve that covers a hollow
tube or bobbin.  The sleeve is impregnated with a
lithium chloride solution (Figure 4.5.3).  Low-voltage
a.c. is supplied to the electrodes, which are not
Interconnected but depend on the conductivity of the
atmospherically moistened lithium chloride for current
flow.   Tne temperature sensor in the tube is usually a
resistance thermometer, but can be a thermistor,
thermocouple, bimetal thermometer, capillary system, or
any sensor calibrated for the proper temperature-to-
dew-point relationship.

       In the early 1960's, the technio^ie of detecting
the dew point on a cooled mirror surface evolved into a
product ion-type unit.  This unit was automatically
operated and had an optical dew-sensing system that
incorporated thermoelectric cooling  (Francisco and
Beaubien, 1963),  Four manufacturers now produce a
aeteorological type, thermoelectric, cooled-mlrror dew
point  instrument (Mazzarella, 1977).  Three of these
instruments cover the range of -50 to +50 C.  Linear
thermistors are used to measure the mirror temperature
in three of the units; a platinum wire sensor is used
ia the other.  All are designed with simultaneous
linear output signals for T
and T (ambient temperature).
(dew point temperature)

Two of the manufacturers
Figure 4.5.3  A
 typical Dewcel
 sensor housing
and transmitter

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    4  of 10

•ake claim to NBS-traeeability with stated dew point accuracies ranging from
±0.2  to ±0.4 C and ambient temperature accuracies ranging from ±0.1  to
±0.5 C.   Ail incorporate some form of standardization that involves clearing
the mirror by heating, either automatically or manually.  Although complex in
design and operation, this type of cooled-mirror hygrometer is considered to be
a functional standard.

       In recent years, two other sensors for humidity have been used on tower
installations for atmospheric pollution studies.  One involves the use of an
organic seed, cut and coupled to a. strain gage.  In principle, absorption of
•oisture in the seed results in distortion, which is converted to an electrical
signal by the strain gage assembly.  Reports on performance are mixed.
Certainly the applications are limited, and the approach does not represent a
technological advance.  By contrast, the thin film capacitor, designed primari-
ly for radiosonde applications, Incorporates advanced technology  (Suntola and
Aotson,  1973).  Reports of users in the past have been mixed, with a common
conplaint of poor performance in polluted atmospheres.  Modern capacitor-type
sensors have achieved a better performance through improved design and user
education.

       4.5.1.1   Sensor Characteristics

       Although the psychrometer is considered the most practical and widely
used instrument for measuring humidity, two major problems are associated with
wet and dry bulb psychrometry involving the accuracy of the thermometers and
the cumulative errors related to operating technique  (Quinn, 1968).  An accur-
acy of ±1 percent at 23°C and SO percent RH requires thermometers with relative
accuracy of ±0.1 C.  The commonly used 0.5 C division thermometers introduce an
uncertainty of ±5 percent RH at this condition.  This assumes that the readings
were taken at the maximum wet bulb depression, a difficult task with a sling
psychrometer.

       It has long been recognized that there are some limitations in using the
dew cell instrument  (Acheson, 1963).  The  lowest relative humidity it can meas-
ure at a given temperature is the ratio of the vapor pressure of  a saturated
solution of LiCl to that of pure water.  This is calculated to be 11.8 percent
RH.  A second limitation is that at -65.6*, -20.5*, +19.0°, and +94.0°C, LiCl
in equilibrium with its saturated solution undergoes a phase change.  Errors in
dew point measurements occur at -69 » -39  , -12 , and +40 C.  This problem is
inherent in the use of LiCl and cannot be eliminated.  It is estimated  that the
                                                    o                       o
accuracy of the LiCl saturated salt technique is 1.5 C over the range of -30
to 30°C.

       The optical chilled (cooled) mirror technique of measuring dew point is
a fundamental measurement.  No calibration is required for the fundamental dew
generating process.  The measurement however is the temperature of the surface
at which the dew forms and as with any electrical temperature measurement
system, calibration is required.  The process of periodically heating the
mirror to a temperature above the dew  (or  frost) point is followed by zeroing
the optical system to correct for the dry  mirror reflectance changing due to
contamination.  In the better instruments, automatic zeroing is programmable in
terms of frequency and length of time.  It can also be accomplished manually.

-------
                                                             Section No.  4.5,0
                                                             Revision No,     0
                                                             Date:     6 Aug 89
                                                             Page:     5  of 10
       4.5.1.2   Sensor Housings and Shields

       Psychrometers of all types should be acclimated to the environmental
conditions in which the measurements are to be made.   In most cases,  psychro-
meters should be stored in a standard instrument shelter so that the mass of
the thermometers,  especially the mass of the housing,  adjusts to the tempei—
ature of the air.   Psychrometers with a stored water supply,  such as those on a
tower, must be shielded from solar radiation.

       For meteorological applications, the dew cell element should be enclosed
in a weatherhood to protect it from precipitation,  wind, and radiation effects.
This type of element functions best in still air.  Some aspirated radiation
shields are designed,  in keeping with these specifications to house both a
temperature sensor, which requires ventilation,  and a dew cell, which requires
only the smallest amount of air flow (Figure 4.5.4).   The miniaturization of
the dew cell has created some problems related to excessive air flow and solar
radiation that remain only partially solved.

       All manufacturers of optical cooled-mirror dew point and temperature
monitoring equipment provide housings for the sensors, which include forced
ventilation and shielding from solar radiation.

       4.5.1.3   Data Requirements

       Electrical hygrometers for monitoring applications have time constants
generally longer than air temperature systems.   The usual data of interest are
hourly average values.  Data should be reported in terms of the condition
                                       measured,  dew point temperature,
                                       relative humidity or wet-bulb and dry-
                                       bulb temperature.  Programs may be used
                                       which convert among these if all the
                                       relevant variables are known.   The
                                       station elevation may be used to
                                       estimate a nominal pressure if a
                                       measurement is not available.   The temp-
                                       erature needed to convert a relative
                                       humidity measurement to dew point temp-
                                       erature is that temperature at the
                                       relative humidity sensor surface.   This
                                       may not be the same temperature as that
                                       measured at some other location.   On the
                                       other hand,  the dew point temperature is
                                       a fundamental measure of the amount of
                                       water vapor in the air and is indepen-
                                       dent of air temperature.  Relative
                                       humidity calculations can therefore be
                                       made given the dew point temperature and
                                       any temperature measurement point in the
Figure 4.5.4  A pair of towei—
mounted Gill aspirated radiation
shields for housing temperature
and dew point sensors (Young),
same general air mass.   Empirical formulae for the estimation of relative
humidity as a function of dew point temperature and air temperature,  relative
humidity as a function of wet-,  dry-bulb temperature and pressure,  and dew
point temperature as a function of relative humidity and temperature are shown
below.

-------
                                                             Section No. 4.5.0
                                                             Revision No.     0
                                                             Date:     6 Aug 89
                                                             Page:    6  of 10
       To calculate relative humidity (RH « 100 r,  %)  from air temperature  {T,
°C)  and dew point temperature (T ,   C),  do the following:
                r = exp
                              b * T
                 b*T.
                                   (1)
                     where a »  17.27
                           b - 237.3
       To calculate the dew point temperature (T ,   C)  from air temperature (T,
    and relative humidity (RH = 100 r, "/,) use
                       T  =
                        D
                                  In r
                  aT
                 b * T
                             a -
                                  In r +
                                           aT
                                   (2)
       To calculate relative humidity (RH - 100 r,  %)  from air temperature
(T,   C),  wet-bulb temperature (T ,   C),  and atmospheric pressure (P, mb)
through the vapor pressure (e» mb)  and the saturation  vapor pressure (e ,  mb),
do the following:
                r =
      where A = 6.6  x 10
            B = 1.15 x 10
          e   =  e  exp
           SH     O   *
                         -4
-3
                            aT
                                  (3)

       To estimate wet-bulb temperature (T ,   C)  from air temperature (T, °C),
dev point temperature (T ,   C), relative humidity (r,  ratio) and atmospheric
pressure (P,   mb), do the following:
T tt T
w
P + T
D
P + T
19 + 130r
19 + 130r
- 28r2 ) "
- 28r2 j
                                                           (4)
The summarization of these relationships was suggested by A.  L.  Morris from
material found in Z. Geophysik,  6, 297, 1930,  the Smithsonian Meteorological
Tables,  Sixth Revised Edition, the Glossary of Meteorology and has been
augmented by his own derivation of expression (4).

-------
                                                             Section No, 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    7  of 10

       Psychrometers are convenient devices for making spot checks of the per-
formance of other devices, especially those that are permanently  installed,
providing the checking is done tinder reasonably steady overcast conditions.
The psyehroiaetric technique built into tower installations presents servicing
problems, especially at temperature extremes.  High temperatures  cause rapid
evaporation, and low temperatures cause freezing.

       Both the dew cell and the cooled-mirror type instruments have applica-
tions on lO-meter or taller tower installations for pollution studies,
providing the sensors are housed in the recommended shields with  little, if
any, aspiration for the dew cell and the recommended rate of aspiration for the
cooled-airror design is selected.

4.5.2  SPECIFICATIONS

       4.5.2.1   Procurement

       The selection of a humidity instrument is guided by the application to
which the data will be put.  The PSD (Prevention of Significant Deterioration)
guideline (EPA, 1987) provides the following:  "...If the permit  granting auth-
ority determines that a significant potential exists for fog formation, icing,
etc., due to effluents from the proposed facility, error in the selected
measurement technique should not exceed an equivalent dewpoint temperature
error of 0.5 C.  Otherwise, errors In equivalent dew-point temperature should
not exceed 1.5°C over a dewpoint range of -30°C to +30°C."  This  latter toler-
ance allows for the use of lithium chloride dew cells.

       Sling psychrometers and aspirated psychrometers with thermometers
shorter than 10 inches do not have sufficient resolution for the  accuracies
required for checking other instruments.—Equally important, the  thermometers
should have etched stems; i.e., the scale markings should be etched on the
glass.  Reliable thermometers are factory calibrated at a minimum of two temp-
eratures, and usually at three.  Thermometers calibrated with NBS-traceable
standards are preferred.

       When patents expired on the original Dewcell, a number of  similar units
appeared on the market.  In light of problems which have existed  in the past,
it is prudent to specify accuracy of the humidity system when it  is operating
as a syste» in the atmosphere.  Problems with ventilation rates will be quickly
exposed by this requirement.  It is not recommended to purchase components to
patch together in a system.  Corrosion in polluted atmospheres can be avoided
by selecting optional 24-carat gold windings, provided cost is not prohibitive.
If dew point alone is to be measured, the standard weatherhood is a proper
choice.  If both temperature and dew point are to be measured, it may be
advantageous to purchase a standard shield that provides a housing for the dew
cell and a separate aspirated compartment for the temperature probe.

       Optical cooled-mirror dew point systems are now commercially available
from several manufacturers, all of which incorporate either linear thermistors
or platinum resistance temperature devices.

-------
                                                             Section No. 4,5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    8  of 10
       4.5.2.2   Acceptance Testing
       Test at least the ambient atmosphere at one point in normal wind and
radiation.

       4.5.2.3   Calibration

       The procedure for calibrating the thermometers in a psychrometer is
essentially the same as any thermometer calibration (See Section 4.3.5).

       Both the dew cell and the cooled-mirror hygrometer can be checked for
approximate calibration accuracy with a motor-operated psychrameter.  Their
performance should be verified under stable conditions at night or under cloudy
conditions during the day.  Several readings taken at the intake of the aspir-
ator or shield are recommended.  Bench calibrations of these more sophisticated
units must be made by the manufacturer.  The electronics portion of some
instruments may be calibrated by substitution of known resistances in place of
the temperature sensor.  This procedure, if appropriate, is described in the
•anufacturer' s operating manual for the instrument.

4.5.3  OPERATIONS

       4.5.3.1   Installation

       Dew point measuring equipment on a  tower should be installed with the
sane considerations given to temperature sensors.  Reference has already been
•ade to the weatherhood as a shield for the dew cell and to an aspirated shield
for the coded-mirror instrument.  At some installations, success has been
reported  in mounting these housings so that they are close to the tower frame-
work on the north-facing side.  This minimizes the effects of direct solar
radiation and provides a rigid support, especially for the cooled-mirror
sensor, wnich requires a stable mounting surface.  Another consideration in
•ounting  these devices inboard involves servicing.  Inboard mounting makes
recharging the dew cell with lithium chloride and cleaning the reflective
surface of the cooled- mirror hygrometer much easier,

       4.5.3.2   Field Operation and Preventive Maintenance

       Field calibration checks should be  made at least monthly on dew cell
type units.  The use of gold wire windings around the L1C1 cylinder minimizes
corrosion problems in polluted atmospheres.  Periodic removal and washing of
old lithium chloride, followed by recharging with a fresh solution, Improves
data reliability.

       Once a mercury or alcohol llquid-in-glass thermometer is calibrated,
there is  no need for recalibration, unless it is to be used for reference or as
a transfer standard.  Errors in wet bulb temperatures are most frequently the
result of an improperly installed or dirty muslin wick, the repeated use of tap
water Instead of distilled water, or human error in reading.  Wicking material
used on psychrometers must be washed to remove traces of sizing and finger-
prints.   Once cleaned, the material is tied at the top of the thermometer bulb
and a loop of thread placed around the bottom so the thermometer bulb is
tightly covered.  To prevent solid materials from collecting on the cloth and
preventing proper evaporation, the wick should be wet with distilled water.  Of

-------
                                                             Section No. 4,5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    9  of 10


course,  slinging or motor aspiration should be done in the shade, away from
reflected or scattered radiation, at a ventilation rate of about 3 to 5 m/s.
Many technique-related errors are minimized by using an Assmann-type, motoi—
operated psychroraeter, providing the instrument Is allowed to assume near
aabient  conditions prior to use,

      The cooled-mirror instruments require no calibration except for the
•inor temperature sensor.  Depending on environmental conditions, the mirror is
easily cleaned with a Q-Tip dipped in the recommended cleaning fluid, usually a
liquid with an alcohol base.  While the accuracy of a psychrometer is inferior
to that  of the optical chilled mirror system, an occasional check at the intake
to the sensor shield is recommended under the provisions specified earlier.

      All operational and preventive maintenance activities should be logged.
Data retrieval will be dependent upon program objectives.

4.S.4 PERFORMANCE AUDIT METHODS

       Instrument audit procedures for hygrometry systems follow calibration
procedures.  A systems audit should be performed near the beginning of a field
  isurement program.
       The performance audit of a humidity measuring system should be based
upon a. comparison with a collocated transfer standard (CTS),  Parts of the
system can be tested by conventional electronic tests, but this avoids so much
of the measurement process that it should only be used to augment the total
system test.  The CTS may be any qualified instrument.  The most accurate type
is the cooled-mirror dew point instrument.  The Assmann-type psychrometer with
calibrated thermometers traceable to NBS is acceptable for most data applica-
tions.  It Is also most convenient since it does not require commercial power
and can be carried to elevated levels on a tower.

       Given the qualifier that humidity is a very difficult measurement to
nake, a rule of thumb for judging the accuracy of a humidity monitoring system
with an Assmann as the CTS is as follows:  when the CTS and the challenged
system agree in dew point temperature to within 1 C, the challenged system is
assumed to be within 0.5°C of the true value;  This arbitrarily assigns an
uncertainty in dew point temperature of +0.5 C for the Assmann which Is true
for most of the range.

       Auditing is best backed by authoritative standards.  ASTM, 1982, 1983,
1984 and 1985 may be of selective value.

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:   10  of 10
4.5.5  REFERENCES
Acheson,  D- T., 1963:  Some limitations and errors inherent in  the use of
    dew-cell for measurement of atmospheric dew points. Monthly Weather Review.
    ASTM,  1982: Standard Definitions of Terms Relating to Humidity
    Measurements, D4023-S2a, American Society for Testing and   Materials,
    Philadelphia, PA.

ASTM, 1983:  Standard Method of Measuring Humidity with Coded-Surface
    Condensation (Dev Point) Hygrometer, D423Q-83, American Society  for
    Testing and  Materials, Philadelphia, PA.

ASTM, 1984:  Standard Test Method for Measuring Humidity with a Psychrometer
    (the Measurement of Vet- and Dry-Bulb Temperatures), E337-84, American
    Society for Testing and  Materials,  Philadelphia, PA.

ASTM, 1985:  Standard Practice for Maintaining Constant Relative  Humidity by
    Means of Aqueous Solutions, E104-85, American Society  for  Testing and
    Materials, Philadelphia, PA.

Berry, F,  A., Jr., E. Bollay and N. R. Beers, 1945:  Handbook of Meteorology.
    McGraw-Hill Book Company, Inc.

Conover,  J. H., 1950:  Tests and adaptation of the Foxboro dew-point  recorder
    for weather observatory use. Bulletin of American  Meteorological
    Society,31(11,13-22.

EPA, 1987:  Ambient Monitoring Guidelines for Prevention of Significant
    Deterioration (PSD). EPA-450/4-87-007, Office of Air Quality  Planning and
    Standards, Research Triangle Park, NC.

Francisco and Beaubien, 1965:  An automatic dew point  hygrometer with
    thermoelectric cooling. Humidity and Moisture, edited by A.Wexler,
    Relnhold Publishing Company.

Hicks, W. F.,  1947:  Humidity measurement by a new system. Refrigerating
    Engineering. American Society of Refrigerating Engineering.

Huschke, R.,ed., 1959:  The Glossary of Meteorology. American   Meteorological
    Society, Boston, MA.

Mazzarella, D. A., 1972:  Meteorological instruments:  their selection and  use
    in air pollution studies. Proceedings of the Meeting on  Education and
    Training  in Meteorological Aspects of Atmospheric  Pollution and Related
    Environmental Problems. World  Meteorological Organization, No. 493.

Middleton, W.  E. K., and A. F. Spilhaus, 1953:  Meteorological  Instruments,
    University of Toronto Press.

Quinn, F. C.,  1963:  Humidity-the neglected parameter. Testing  Engineering,
    The Mattingly Publishing Company,  Inc.

Suntola and Antson,  1973:  A thin film humidity sensor. Scientific
    Discussions, CIMO VI, World Meteorological Organization.

-------
                                                             Section No. 4.6.0
                                                             Revision No.     0
                                                             Date:   17 Sep  89
                                                             Page:     1   of 8
              QUALITY ASSURANCE FOR SOLAR RADIATION MEASUREMENTS

4.6.0  INTRODUCTION

    Solar energy is the driving force of  large-scale atmospheric motion,
indeed,  of the general circulation of the atmosphere.  Although air pollution
investigators normally consider the measurement of solar radiation secondary to
wind and temperature measurements, solar  radiation is directly related  to
atmospheric stability.   It is measured as total Incoming global radiation,  as
outgoing reflected and terrestrial radiation and as net total radiation.

    Quantitatively, solar radiation is described in units of energy flux,
either W/ra2 or cal/cm *min.   When measured in specific,  narrow wavelength
bands, solar radiation may be used to evaluate such air pollution  indicators as
turbidity, amount of precipitable water,  and rates of photochemical reactions.
However, this manual will cover only broadband measurements and sunshine.

    The generic term, radiometer, refers  to any instrument that measures
radiation, regardless of wavelength.  Shortwave radiation has wavelengths  less
than 4 micrometers (nm) and is subdivided as follows:
                                        0.20 pro to 0.38 pn
                                        0.38 pm to 0.75 JJTO
                                        0.75 urn to 4.00 fun
                      Ultraviolet CUV)
                      Visible
                      Near-infrared
    Longwave radiation has a wavelength as follows:
                      Infrared (IR)     4    pm to  100 urn

and comes from the Earth and its atmosphere.  The  instruments  most  commonly
used for environmental monitoring  are discussed below.

4.6.1  TYPES OF INSTRUMENTS

       4.6.1.1 Pyranometers

       Pyranometers are instruments  that measure the  solar  radiation  received
fro» the hemispherical part of
the atmosphere it sees,
including the total sun and
                                            Sensing
sky shortwave radiation on a
horizontal surface (Figure
4.6.1).   Most pyranometers
incorporate a thermopile as
sensor.   Some use a silicon
photovoltaic cell as a sensor.
The precision spectral
pyranometer (PSP) is made by
Eppley Laboratories and has
two hemispherical domes
designed to measure sun and
sky radiation on a horizontal
surface in defined
wavelengths.  This is achieved
by substituting one of several
                                                               y Precision Ground &
                                             Element    ^-^—~^>y Polished Glass Dome
                                                                (Frequently Double}
                                                                         Leveling
                                                                         Screw
                                 Figure 4.6.1
Features of a typical
pyranometer (Carter, et al.,1977)

-------
                                                             Section No, 4.6,0
                                                             Revision No,    0
                                                             Date:   17 Sep 89
                                                             Page:     2  of 8

colored Schott glass filter domes for the clear glass outer dome.  The smaller
doae suppresses convection, so this type is better sited if tilted from the
horizontal.

       4.6.1.2 Bimetallic Recording Pyranoneters

       Bimetallic recording pyranometers, also known as actln- oneters,
predesigned by Robitzsch of Germany,  these mechanical sensors consist of two
or three bimetallic strips, alternately painted black and white, that
respectively absorb and reflect solar radiation.  The resulting differential
heating produces a deformation that is transmitted mechanically through levers
and a pen arm to a clock-wound drum recorder.  Although of limited accuracy,
these instruments are useful for locations with no commercial power.

       4.6.1.3 Net Radiometers

       Net radiometers or net pyrradlometers are designed to measure the
difference between downward and upward total radiation, including the total .
incoming shortwave and longwave radiation and the
total outgoing shortwave and longwave radiation.
There are two basic types of net radiometers.  The
ventilated plate type, often referred to by the
nane of the designers (Gler and Dunkle), Is more
popular in research applications than the type
with hemispherical polyethylene domes originally
designed by Funk.  Both incorporate thermopiles
with blackened surfaces.  Because net radiometers
produce a signal with a. positive sign when the
incoming radiation exceeds the outgoing, the
recording equipment must be designed with an
offset zero.

       4.6.1.4 Sunshine Recorders

       Sunshine recorders are designed to provide
information on the hourly or daily duration of
sunshine.  Only one commercially available,
off-the-shelf type of sunshine recorder is now
available.  This is the Campbell-Stokes design
(Figure 4.6.2), designated as the Interim
reference sunshine recorder "IRSR" by the World
Meteorological Organization.  The device consists
of a glass sphere 10 cm in diameter mounted in a
spherical bowl.  The sun's rays are focused on a
card that absorbs radiation and changes color in
the presence of sunlight.  The recorder is used
infrequently In the United States but extensive-
ly abroad, primarily for the collection of
climatological data.  The National Weather Service
routinely uses a Sunshine Switch, which             Figure 4.6.2  A Campbell-
incorporates one shaded photocell and one exposed   Stokes Sunshine Recorder
photocell.                                          (U.S. Army, 1975)

-------
                                                            Section No.  4.6.0
                                                            Revision No.     0
                                                            Date:   17 Sep 89
                                                            Page;     3   of 8
Declination Adjustment

       24 hf. Dial
       4.6.1.5 Instrument Characteristics

       Only the characteristics of pyranometers and net  radiometers,  the two
types of Instruments used most frequently in pollution-related programs, will
be discussed in this section.   The pyranoroeter is not to be confused with the
pyrhellometer, "an instrument  for measuring the Intensity of direct solar
                                   , ,     radiation at normal Incidence*  (WHO,
                                         1971).  The pyrhellometer is mounted
                                         in a solar tracker, or equatorial
                                         mount, automatically pointing to the
                                         sun as it traverses from east to west
                                         (Figure 4.6.3).  By contrast, the
                                         pyranoBieter Is  mounted facing toward
                                         the zenith.  Ideally, the response of
                                         the .thermopile  sensor in the
                                         pyranometer ^s  proportional to the
                                         cosine of the angle of the solar beam
                                         and is constant at all azimuth angles.
                                         This characteristic is known as the
                                         Lambert Cosine  Response, an Important
                                         characteristic  of pyranometers.  For
                                         the majority of applications related
                                         to atmospheric  pollution, Class 2 and
                                         Class 3 are satisfactory (see Table
Figure 4.6,3  Features of &              4.6.1),
typical pyrheliometer and track-
ing count (Carter, et a.l.f 1977)                 Most net radiometers now
                                         available commercially are made with a
small disc-shaped thermopile covered by polyethylene hemispheres.  In most
units the material used for shielding the element from the wind and weather  is
very thin and is transparent to wavelengths of 0.3 to 60 fjnu  Until recently,
the internal ventilation and positive pressure required to maintain the
hemispheres of net radiometers in their proper shape was considered critical;
however, new designs have eliminated this problem.  The plate-type net
radiometer, most often the modified Gier and Dunkle design sold commercially in
the United States, is occasionally used in routine air pollution
investigations.  The thermopile heat flow transducer is blackened with a
material that is easily cleaned with water or naphtha.  Because the thermopile
is uncovered for total spectrum response, a built-in blower, available for
operation on 115 V 50/60 Hz or 12 V d.c., draws air across the element at a
constant rate eliminating the effects of varying natural winds.  The device  is
temper a ture-compensa ted and typically has a sensitivity of 2.2 fiV per W/m ,  a
response time of 10 seconds, and a "relative" accuracy of two percent in
calibration.  When supplied with a reflective shield on its lower surface, this
plate type net radiometer of the Gier and Dunkle design becomes a total
hemispherical radiometer or unshielded pyranometer.

       4.6.1.6  Recorders and Integrators for Pyranoaeters and
                Net Radiometers

       The relatively high impedance and low signal of thermopile sensors,
excluding silicon photovoltaic cells, limits their use with both Indicating
meters and recording meters.  Electronic strip chart millivolt potentiometric
recorders incorporating variable-range rheostats are preferred.  The

-------
                                                             Section No. 4,6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     4  of 8
variable-range rheostat permits the exact matching of the recorder scale to
interchangeable sensors so that deflections of the meter represent engineering
units,  i.e..  W/m2, cal/cmz»min, etc.   The alternative is a standard millivolt-
meter potentiometrlc recorder where the data,  in millivolts,  must be translated
to units of energy, corresponding to full-scale values of 1370 W/m  or 1.96
cal/cm  «ain.   It may also be necessary, especially if the signal is to be used
as an input to a computer, to combine preaaplificatlon with scaling.

4.6.2  SPECIFICATIONS

       4.6.2.1   Procurement

       In purchasing a solar radiation measurement system, follow the practice
of aatching the data requirements to the Instrument selection, specifying the
performance required on the purchase order (complete with test method to verify
performance) and testing the performance in receiving.  See Section 4.1,4.5 for
co«*ents on traceability protocol.  Many types of radiation instruments have
been developed, especially in recent years, because of an increasing interest
in environmental considerations (Gates, 1962), meteorological research
(Monteith, 1972), and solar energy (Carter, et al., 1977).  Except for special
studies, the retirements for relating radiation to stability can be satisfied
by purchasing sensors of Class 2 or Class 3 as identified by the WHO (see Table
4.6.1).

       Table 4.6.1 Classification of Pyranometers According to
                        Physical Response Characteristics
Sens.
(mW/cm2)
1st Class ±0.1
2nd Class ±0.5
3rd Class ±1.0
Temp.
± 1
± 2
± 5
Lin.
( % 3
± 1
± 2
± 3
Max Time
Constant
(sec. )
25
60
240
Cosine
Response
± 3.0
± 5.7
± 10
       Class 2 sensors offer the advantage of providing data comparable to  that
collected at National Weather Service stations and at key locations of
Department of Energy (DOE). The sensors to be specified should be commercially
available, field proven by the Manufacturer for several years, and have the
technical requirements established by WHO standards.  Several American Society
for Testing and Materials  (ASTM) standards are available (ASTM,  1984).  When
purchasing a recorder or integrator, one must match the calibration factor  or
sensitivity of the sensor  to the readout equipment.  It must be  recognized  that
the signals from net radiometers, in contrast to pyranometers, require
zero-offset capability to accommodate both negative and positive voltage
outputs.

       4.6.2.2 Acceptance Test Ing

       Physical Inspection of the relatively fragile pyranometers or  net
radiometers immediately after delivery of the instrument is important.  One
must be sure that the calibration data have been received and that these  data
correspond to the serial number of the instrument.  Storage of this information

-------
                                                             Section No. 4.6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     5  of 8


will prove helpful when the time comes to have the calibration of the
instrument checked, or to replace the sensor or readout device.  Few
organizations are equipped or staffed to bench-test a radiometer to verify
calibration, but a quick determination can be made indoors as to whether the
sensor and recorder or integrator system is operating by exposing the sensor to
the light of a tungsten lamp.  It may be necessary to place the instrument
fairly close to the lamp.  Covering the sensor for several hours will ensure
that the system is not "dark counting."

       4.6.2.3  Calibration

       The user of a pyrano»eter or net radiometer is normally not equipped to
calibrate the sensor.  The best the user can do is to perform field calibration
checks on two cloudless sky days.  These checks involve a side-by-side
comparison of the sensor to a sensor of similar design, the calibration of
which can be traced to a transfer standard.  Since 1975 all measurements have
been made in accordance with the Absplute Radiation Scale or equivalently the
World Radiometric Reference established at the International Pyrhe Home trie
Comparison IV at Davos, Switzerland {NCAR, 1984,  pp. 4-103).  If a side-by-side
comparison is not possible, the device should be returned to the manufacturer
or to a laboratory with the facilities to check the calibration.  The frequency
of making comparative readings or having factory calibrations will depend on
environmental conditions.  Any indication of discoloration or peeling of a
blackened surface or of scratches on the hemispheres of a pyranometer warrants
recalibration and/or service.

          Net radiometers are more delicate and require more frequent attention
than pyranometers.  Pyranometers of high quality in a clean atmosphere may
require recalibration annually; net radiometers should be recalibrated at least
yearly.  Calibrating the recorder or integrator is an easy task.  The standard
method involves the use of a precision potentiometer to impress known voltages
into the circuit.  The linearity of the readout instrument may be checked by
introducing a series of voltages covering the full scale, checking first
up-scale and then down-scale.  Adjustments should be made as necessary.  In the
absence of a precision potentiometer, it may be possible to introduce a
calibrated millivolt source that covers one or two points.  Integrators can be
checked the same way, except that the Input value must also be timed.

4.6.3  OPERATIONS

       4.6.3.1 Installation

       The site selected for an upward-looking pyranometer should be free from
any obstruction above the plane of the sensor and should be readily accessible
for cleaning and maintenance.  It should be located so that shadows will not be
cast on the device, and away from light-colored walls or other objects likely
to reflect sunlight.  A flat roof is usually a good choice; but if such a site
is not possible, a rigid stand with a horizontal surface some distance from
buildings or other obstructions should be used.  A site survey of the angular
elevation above the plane of the radiometer surface should be made through 360
degrees (The Eppley Laboratory, Inc.),

       The same procedures and precautions should be followed for net
radiometers that are both upward- and downward-looking.  However, the

-------
                                                             Section No.  4.6.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:      6  of 8
instrument must be supported on an arm extending from a vertical support about
1 • above the ground.   Except for net radiometers with heavy-duty domes, which
are installed with a desiccant tube in series with the sensor chamber,  most
other hemispherical net radiometers require the positive pressure of a
gas—usually nitrogen—to both maintain the shape of the polyethylene domes and
purge the area surrounding the thermopile.  In on increasingly popular design,
there is a requirement for internal purging with nitrogen and external
ventilation with compressed dry air through holes on the frame.  The compressed
air supply minimizes fogging and condensation.

       Precautions must be taken
to avoid subjecting radiometers      ;              .
to mechanical shock during
installation.  They should be
installed securely and leveled
using the circular spirit level
attached to the instrument.  Net
radiometers are difficult to
•ount and to maintain free of
vibration.  Pyranometers of the
Holl-Gorzynski design, used
extensively by Atmospheric
Environmental Sciences (AES) of
Canada,  are oriented so that the
emerging leads face north (Figure
4.6.4).   This minimizes solar
heat on the electrical
connections of an instrument that
is not temperature compensated.
The thermopiles of these
instruments should be oriented so
that the long side of the
         «  10  in.
Figure 4.6.4  A Moll-Gorczynskla
   Solarimeter (U.S. Army, 1975)
thermopile points east and west (Latimer, 1972).  The cable used to connect the
pyranometer to the readout device, recorder, or integrator should be between
16 and 20 gauge and made of shielded, waterproofed 2-conductor copper wire.
The sensor, shield, and readout device should be connected to a common ground.
Potentiometrie millivolt recorders are to be used with most high-impedance,
low-signal radiometers.  Cable lengths of 300 m or more are practical.
Galvanometrlc recorders can be used with silicon cell radiometers.  Soldered,
copper-to-copper junctions between instrument connectors and/or cables are
essential.  Pyranographs or actlnographs should be installed on a level surface
immune to shadows.  These instruments should be placed in such a way that the
sensitive bimetallic strips lie within 2 degrees of true east and west with the
glass inspection window facing north (in the northern hemisphere).

       4.6.3.2 Field Operation of a Solar Radiation System

       As part of the quality assurance program, a field calibration check
should be performed at least once every 6 months according to the procedures
outlined in Section 4.6.2.3.  Solar radiation instruments require almost daily
attention.  The data should be inspected for a reasonable diurnal pattern and
the absence of dark counting.  Where strip chart or digital printers are used,
daily time checks are desirable.  Data retrieval will depead upon program

-------
                                                             Section No. 4.6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     7  of 8

objectives; but even for climatological programs, data should be collected
monthly.   All operational activities during a site visit should be logged.

       4.6.3.3 Preventive Maintenance

       All types of radiometers require frequent cleaning to remove any
material deposited on the surface that will intercept the radiation.  In most
cases,  this is a dally operation.  The outer hemisphere should be wiped clean
and dry with a lint-free soft cloth, using alcohol..   Any scratching of the
surface will alter the transmission properties of the glass, so cleaning must
be done with care.  If frozen snow, glazed ice, hoarfrost or rime ice is
present,  an attempt should be made to remove the deposit carefully with warmed
cloths.

       Should the internal surface of a pyranometer' s outer hemisphere become
coated with moisture, It can be cleaned by carefully removing the outer
hemisphere on a dry day and allowing the air to evaporate the moisture, then
checking the dessicant.  If removal of a hemisphere exposes the thermopile
element,  extreme care should be taken because it is fragile and easily damaged.
About once each month, the desiccant installed in most pyranometers should be
inspected.  Whenever the silica gel drying agent is pink or white instead of
blue, it should be replaced or rejuvenated by drying it out on a pan in 135°C
oven.  The level should be checked after each servicing of the radiometer, or
at least Monthly.  Significant errors can result from misalignment.

       Net radiometers require more frequent maintenance attention than
pyranometers.  It is necessary to replace the polyethylene domes as often as
twice a year or more before the domes become discolored, distorted, or cracked.
More frequent replacement is necessary in polluted environments due to
accelerated degradation of plastic hemispheres when exposed to pollutants.  A
daily maintenance schedule is essential to check on the proper flow of gas in
instruments that are inflated and purged with nitrogen.   All PM activities
should be recorded in a log.

4.6.4  PERFORMANCE AUDIT METHODS

    A performance audit on a solar radiation system is only practical with a
CTS.  The CTS must have the spectral response and exposure equivalent to the
instrument being audited.  One diurnal cycle will establish an estimate of
accuracy sufficient for most air quality monitoring applications.  The method
of reporting the data from the monitoring instrument (daily integrated value,
hourly integrated value, average intensity per hour,  etc.) must be used in
reducing the data from the CTS to provide a meaningful comparison.  An audit
frequency of at least six months is recommended.

4.6.5  REFERENCES

ASTM, 1984:  Calibration of secondary reference pyrhellometers and
    pyrheliometers for field use, E816. American Society for Testing and
    Materials, Philadelphia, PA.

Carter,  E. A. et al., 1977:  Catalog of solar radiation measuring equipment.
    ERDA/ORO/5361-l,  U.S. Energy and Development Administration.

-------
                                                              Section No, 4.6.0
                                                              Revision No.     0
                                                              Date:   17 Sep  89
                                                              Page:     8  of 8

Gates,  D. ML ,  1962:  Energy Exchange in the Biosphere,  Harper and Row.

Latimer, J. R., 1972:  Radiation measurement.  Technical Manual Series No. 2,
    International Field Year  for the Great Lakes, Canadian National Commission
    for the Hydrologlcal Decade.

Monteith, J. L., 1972:  Survey of  instruments for micrometeorology.
    Internalional Biological  Programs ffandfeooJt No. 22. Blackwell Scientific
    Publications, Osney Mead,  Oxford,  England.

NCAR,  1984:  Instructor's  Handbook on Meteorological Instrumentation,
    F.  V. Brock, Editor. NCAR Technical Mot*,  NCAR/TN-237+1A.

U.S.Army, 1975;  Part 2, natural environmental factors. Engineering Design
    Handbook,  Environmental Series.  Department of the Army, Material Command.

WHO, 1971:  Guide to meteorological instrument and observing practices. World
    Meteorological Organization  No.  8TP3,  4th edition, Geneva', Switzerland.
                        • O.S. COVESRMEirr PRINTING 
-------