EPA Document #: EPA/600/R-14/098 METHOD 543. DETERMINATION OF SELECTED ORGANIC CHEMICALS IN DRINKING WATER BY ON-LINE SOLID PHASE EXTRACTION- LIQUID CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY (ON-LINE SPE-LC/MS/MS) Version 1.0 March 2015 J.A. Shoemaker US EPA, Office of Research and Development, National Exposure Research Laboratory NATIONAL EXPOSURE RESEARCH LABORATORY OFFICE OF RESEARCH AND DEVELOPMENT U. S. ENVIRONMENTAL PROTECTION AGENCY CINCINNATI, OHIO 45268 543-1 ------- METHOD 543 DETERMINATION OF SELECTED ORGANIC CHEMICALS IN DRINKING WATER BY ON-LINE SOLID PHASE EXTRACTION-LIQUID CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY (ON-LINE SPE-LC/MS/MS) 1. SCOPE AND APPLICATION 1.1 This is an on-line solid phase extraction liquid chromatography/tandem mass spectrometry (on-line SPE-LC/MS/MS) method for determination of organic chemicals in drinking water. Accuracy and precision data have been generated in reagent water, and finished ground and surface waters for compounds listed in the table below. This method cannot be used for manual (off-line) SPE extractions. EPA Method 5401 is available for analysis of these analytes by off-line SPE- LC/MS/MS analysis, if desired. Chemical Abstract Services Analvte Registry Number (CASRN) 3-Hydroxycarbofuran 16655-82-6 Bensulide 741-58-2 Fenamiphos 22224-92-6 Fenamiphos sulfone 31972-44-8 Fenamiphos sulfoxide 31972-43-7 Tebuconazole 107534-96-3 Tebufenozide 112410-23-8 1.2 The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets Data Quality Objectives (DQOs) that are developed based on the intended use of this method. The single laboratory lowest concentration MRL (LCMRL) is the lowest true concentration for which the future recovery is predicted to fall, with high confidence (99%), between 50 and 150% recovery. Single laboratory LCMRLs for analytes in this method range from 0.27-1.7 ng/L, and are listed in Table 6. The procedure used to determine the LCMRL is described elsewhere.2 1.3 Laboratories using this method will not be required to determine the LCMRL for this method, but will need to demonstrate that their laboratory MRL meets the require- ments described in Section 9.2.4. 1.4 Determining the Detection Limit (DL) for analytes in this method is optional (Sect. 9.2.6). Detection limit is defined as the statistically calculated minimum concentration that can be measured with 99% confidence that the reported value is greater than zero.3 The DL is compound dependent and is dependent on extraction efficiency, sample 543-2 ------- matrix, fortification concentration, and instrument performance. DLs for analytes in this method range from 0.13-0.99 ng/L, and are listed in Table 6. 1.5 This method is intended for use by analysts skilled in solid phase extractions, operation of LC/MS/MS instruments, and the interpretation of associated data. 1.6 METHOD FLEXIBILITY - In recognition of technological advances in analytical systems and techniques, the laboratory is permitted to modify the analytical LC column, LC gradient and MS and MS/MS conditions (Sect. 6.5.8, 9.1.1, 10.2, and 12.1). Changes may not be made to sample collection and preservation (Sect. 8), sample extraction steps (Sect. 11.3), or to quality control requirements (Sect. 9). Method modifications should be considered only to improve method performance. Modifications that are introduced in the interest of reducing cost or sample processing time, but result in poorer method performance, should not be used. Analvtes must be adequately resolved chromatographically in order to permit the mass spectrometer to dwell on a minimum number of compounds eluting within a retention time window. Instrumental sensitivity (or signal-to-noise) will decrease if too many compounds are permitted to elute within a retention time window. In all cases where method modifications are proposed, the analyst must perform the procedures outlined in the initial demonstration of capability (IDC, Sect. 9.2), verify that all Quality Control (QC) acceptance criteria (Sect. 9) are met, and that acceptable method performance can be verified in a real sample matrix (Sect. 9.3.5). NOTE: The above method flexibility section is intended as an abbreviated summation of method flexibility. Sections 4-12 provide detailed information of specific portions of the method that may be modified. If there is any perceived conflict between the general method flexibility statement in Section 1.6 and specific information in Sections 4-12, Sections 4-12 supersede Section 1.6. 2. SUMMARY OF METHOD A 10-mL preserved water sample is fortified with internal standards and analyzed by automated on-line SPE-LC/MS/MS. A 2-mL aliquot of the sample is loaded onto the in-line SPE cartridge. The SPE cartridge is washed with 20 mM ammonium acetate. Analytes are back-eluted from the solid phase with analytical LC gradient. Analytes are separated and identified by comparing the acquired mass spectra and retention times to reference spectra and retention times for calibration standards acquired under identical LC/MS/MS conditions. The concentration of each analyte is determined by using the internal standard technique. 3. DEFINITIONS 3.1 ANALYSIS BATCH - A set of samples that is analyzed on the same instrument during a 24-hour period, including no more than 20 Field Samples, that begins and ends with the analysis of the appropriate Continuing Calibration Check (CCC) standards. Additional CCCs may be required depending on the length of the analysis batch and/or the number of Field Samples. 543-3 ------- 3.2 CALIBRATION STANDARD (CAL) - A solution prepared from the primary dilution standard solution and/or stock standard solution, and the internal standard(s). The CAL solutions are used to calibrate the instrument response with respect to analyte concentration. 3.3 COLLISIONALLY ACTIVATED DISSOCIATION (CAD) - The process of converting the translational energy of the precursor ion into internal energy by collisions with neutral gas molecules to bring about dissociation into product ions. 3.4 CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing the method analytes and internal standard(s). The CCC is analyzed periodically to verify the accuracy of the existing calibration for those analytes. 3.5 DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be identified, measured, and reported with 99% confidence that the analyte concentration is greater than zero. This is a statistical determination of precision (Sect. 9.2.6), and accurate quantitation is not expected at this level.3 3.6 EXTRACTION BATCH - A set of up to 20 Field Samples (not including QC samples) extracted together by the same person(s) during a work day using the same lot of SPE devices, solvents, internal standard and fortifying solutions. Required QC samples include Laboratory Reagent Blank, Laboratory Fortified Blank, Laboratory Fortified Sample Matrix, and either a Field Duplicate or Laboratory Fortified Sample Matrix Duplicate. 3.7 FIELD DUPLICATES (FD1 and FD2) - Two separate samples collected at the same time and place under identical circumstances, and treated exactly the same throughout field and laboratory procedures. Analyses of FD1 and FD2 give a measure of the precision associated with sample collection, preservation, and storage, as well as laboratory procedures. 3.8 INTERNAL STANDARD (IS) - A pure chemical added to a sample or standard solution in a known amount(s) and used to measure the relative response of other method analytes that are components of the same solution. The internal standard must be a chemical that is structurally similar to the method analytes, has no potential to be present in water samples, and is not a method analyte. 3.9 LABORATORY FORTIFIED BLANK (LFB) - A volume of reagent water or other blank matrix to which known quantities of the method analytes and all the preservation compounds are added in the laboratory. The LFB is analyzed exactly like a sample, and its purpose is to determine whether the methodology is in control, and whether the laboratory is capable of making accurate and precise measurements. 3.10 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - A preserved field sample to which known quantities of the method analytes are added in the laboratory. 543-4 ------- The LFSM is processed and analyzed exactly like a sample, and its purpose is to determine whether the sample matrix contributes bias to the analytical results. The background concentrations of the analytes in the sample matrix must be determined in a separate sample extraction and the measured values in the LFSM corrected for background concentrations. 3.11 LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A duplicate of the Field Sample used to prepare the LFSM. The LFSMD is fortified, extracted, and analyzed identically to the LFSM. The LFSMD is used instead of the Field Duplicate to assess method precision when the occurrence of method analytes is low. 3.12 LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other blank matrix that is treated exactly as a sample including exposure to all glassware, equipment, solvents and reagents, sample preservatives, and internal standards that are used in the analysis batch. The LRB is used to determine if method analytes or other interferences are present in the laboratory environment, the reagents, or the apparatus. 3.13 LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The single laboratory LCMRL is the lowest true concentration for which a future recovery is expected, with 99% confidence, to be between 50 and 150% recovery.2 3.14 MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and reactivity data including storage, spill, and handling precautions. 3.15 MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be reported as a quantitated value for a method analyte in a sample following analysis. This defined concentration can be no lower than the concentration of the lowest calibration standard for that analyte and can only be used if acceptable QC criteria for this standard are met. A procedure for verifying a laboratory's MRL is provided in Section 9.2.4. 3.16 PRECURSOR ION - For the purpose of this method, the precursor ion is the protonated molecule ([M+H]+) of the method analyte. In MS/MS, the precursor ion is mass selected and fragmented by CAD to produce distinctive product ions of smaller 3.17 PRIMARY DILUTION STANDARD (PDS) SOLUTION - A solution containing the analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions. 3.18 PRODUCT ION - For the purpose of this method, a product ion is one of the fragment ions produced in MS/MS by CAD of the precursor ion. 543-5 ------- 3.19 QUALITY CONTROL SAMPLE (QCS) - A solution of method analytes of known concentrations that is obtained from a source external to the laboratory and different from the source of calibration standards. The second source stock standard solution is used to fortify the QCS at a known concentration. The QCS is used to check calibration standard integrity. 3.20 STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing one or more method analytes prepared in the laboratory using assayed reference materials or purchased from a reputable commercial source. 4. INTERFERENCES 4.1 All glassware must be meticulously cleaned. Wash glassware with detergent and tap water, rinse with tap water, followed by a reagent water rinse. Non-volumetric glassware can be heated in a muffle furnace at 400 °C for two hours or solvent rinsed. Volumetric glassware should be solvent rinsed and not be heated in an oven above 120 °C. 4.2 Method interferences may be caused by contaminants in solvents, reagents (including reagent water), sample bottles and caps, and other sample processing hardware that lead to discrete artifacts and/or elevated baselines in chromatograms. All items must be routinely demonstrated to be free from interferences (less than 1/3 the MRL for each method analyte) under the conditions of the analysis by analyzing laboratory reagent blanks as described in Section 9.3.1. Subtracting blank values from sample results is not permitted. 4.3 Matrix interferences may be caused by contaminants that are co-extracted from the sample. The extent of matrix interferences will vary considerably from source to source, depending upon the nature of the water. Humic and/or fulvic material can be co-extracted during SPE and high levels can cause enhancement and/or suppression in the electrospray ionization (ESI) source or low recoveries on the SPE sorbent.4"5 Total organic carbon (TOC) is a good indicator of humic content of the sample. 4.4 Relatively large quantities of the preservatives (Sect. 8.1.2) are added to sample bottles. The potential exists for trace-level organic contaminants in these reagents. Interferences from these sources should be monitored by analysis of laboratory reagent blanks (Sect. 9.3.1), particularly when new lots of reagents are acquired. 4.5 SPE cartridges can be a source of interferences. Analysis of field and laboratory reagent blanks can provide important information regarding the presence or absence of such interferences. Brands and lots of SPE devices should be tested to ensure that contamination does not preclude analyte identification and quantitation. 543-6 ------- 5. SAFETY 5.1 The toxicity or carcinogen!city of each reagent used in this method has not been precisely defined. Each chemical should be treated as a potential health hazard, and exposure to these chemicals should be minimized. Each laboratory is responsible for maintaining an awareness of OSHA regulations regarding safe handling of chemicals used in this method. A reference file of MSDSs should be made available to all personnel involved in the chemical analysis. Additional references to laboratory safety are available.6"8 5.2 Pure standard materials and stock standard solutions of these method analytes should be handled with suitable protection to skin and eyes, and care should be taken not to breathe the vapors or ingest the materials. 6. EQUIPMENT AND SUPPLIES (Brand names and/or catalog numbers are included for illustration only, and do not imply endorsement of the product.) 6.1 SAMPLE CONTAINERS - 100-mL amber glass bottles fitted with polytetrafluoroethylene (PTFE)-lined screw caps. Smaller amber bottles may be used provided the corresponding amount of preservatives can be accurately measured at smaller volumes. 6.2 AUTOSAMPLER VIALS - Amber glass 20-mL autosampler vials (Kimble Chase #60960A-4 or equivalent) with caps containing pre-slit PTFE-silicone septa (Waters #186006322 or equivalent). Smaller volume autosampler vials are also acceptable. 6.3 MICRO SYRINGES - Suggested sizes include 5, 10, 25, 500, 100, 250, 50, 1000, and 10,000-jiL syringes. 6.4 ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g. 6.5 ON-LINE SOLID PHASE EXTRACTION (SPE) LIQUID CHROMATOGRAPHY (LC)/TANDEM MASS SPECTROMETER (MS/MS) WITH DATA SYSTEM (See Figures 2 and 3 for system schematic.) 6.5.1 SPE CARTRIDGES - Waters Oasis HLB, Direct Connect, 20 |im, 2.1x30 mm cartridges (Waters #186005231) - divinylbenzene N-vinylpyrrolidone copolymer. 6.5.2 LOADING (SPE) PUMP - A binary or quaternary LC pump capable of flow rates up to 2 mL/min used to deliver sample loading, cartridge washing and cartridge conditioning solvents. 6.5.3 AUTOSAMPLER - Autosampler must be capable of reproducibly injecting sample volumes up to 2 mL. 543-7 ------- 6.5.4 SWITCHING VALVES - Automatic switching valves capable of switching flows between the loading and analytical pump through the extraction column and analytical column. The two parallel cartridge on-line SPE system used during method development operated with two switching valves. Other configurations that can perform the procedure in Section 11.3 and meet QC criteria in Section 9 are acceptable. 6.5.5 ANALYTCAL LC PUMP - A binary or quaternary LC pump capable of performing binary linear gradients at a constant flow rate near the flow rate used for development of this method (0.4 mL/min). Usage of a cooled autosampler compartment and a column heater is optional. During method development samples were not cooled and the analytical LC column was thermostated at 30 °C. 6.5.6 TANDEM MASS SPECTROMETER - The mass spectrometer must be capable of positive ion electrospray ionization (ESI) near the suggested LC flow rate of 0.4 mL/min. The system must be capable of performing MS/MS to produce unique product ions (Sect. 3.18) for method analytes within specified retention time segments. A minimum of 10 scans across the chromatographic peak is required to ensure adequate precision. Data are demonstrated in Tables 7-10 using a triple quadrupole mass spectrometer. 6.5.7 DATA SYSTEM - An interfaced data system is required to acquire, store, reduce, and output mass spectral data. The computer software should have the capability of processing stored LC/MS/MS data by recognizing an LC peak within any given retention time window. The software must allow integration of the ion abundance of any specific ion within specified time or scan number limits. The software must be able to calculate relative response factors, construct linear regressions or quadratic calibration curves, and calculate analyte concentrations. 6.5.8 ANALYTICAL COLUMN - A UPLC HSS T3 column (2.1 x 50 mm) packed with 1.8 jam Ci8 solid phase particles (Waters #186003538) was used. Any equivalent column that provides adequate resolution, peak shape, capacity, accuracy, and precision (Sect. 9) may be used. CAUTION: Under the sample, SPE, and mobile phase conditions of this method, not all columns performed equivalently. Broad peak shapes were obtained on some Cis columns. A column designed to more strongly retain polar analytes in highly aqueous mobile phases is needed for this method. 6.5.9 IN-LINE FILTER UNIT - A UPLC in-line filter holder (Waters #205000343 or equivalent) with 0.2 |im, 2.1 mm stainless steel filters (Waters #700005698 or equivalent). 543-8 ------- Note: The in-line filter unit is optional, but highly recommended to extend the lifetime of the analytical column from particle buildup at the head of the column. 7. REAGENTS AND STANDARDS 7.1 GASES, REAGENTS, AND SOLVENTS - Reagent grade or better chemicals should be used. Unless otherwise indicated, it is intended that all reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available. Other grades may be used, provided it is first determined that the reagent is of sufficiently high purity to permit its use without lessening the quality of the determination. 7.1.1 REAGENT WATER - Purified water which does not contain any measurable quantities of any method analytes or interfering compounds greater than 1/3 the MRL for each method analyte of interest. 7.1.2 METHANOL (CH3OH, CAS#: 67-56-1) - High purity, demonstrated to be free of analytes and interferences (Fisher Optima LC/MS grade or equivalent). 7.1.3 ACETONITRILE (CH3CN, CAS#: 75-05-8) - High purity, demonstrated to be free of analytes and interferences (Fisher Optima LC/MS grade or equivalent). 7.1.4 AMMONIUM ACETATE (C2H7NO2, CAS# 631 -61 -8) - High purity, demonstrated to be free of analytes and interferences (Sigma LC/MS grade or equivalent). 7.1.5 20 mM AMMONIUM ACETATE - To prepare 1 L, add 1.54 g ammonium acetate to 1 L of reagent water. This solution is prone to volatility losses and should be replaced at least every 48 hours. 7.1.6 SAMPLE PRESERVATION REAGENTS - The following preservatives are solids at room temperature and may be added to the sample bottle before shipment to the field. 7.1.6.1 TRIZMA PRESET CRYSTALS, pH 7.0 (Sigma-Aldrich #T-7193 or equivalent) - Reagent grade. A premixed blend of Tris [Tris(hydroxy- methyl)aminomethane] and Tris HCL [Tris(hydroxymethyl)aminomethane hydrochloride]. Alternatively, a mix of the two components with a weight ratio of 15.5/1 Tris HCL/Tris may be used. These blends are targeted to produce a pH near 7.0 at 25 °C in reagent water. Trizma functions as a buffer (Sect. 8.1.2). 7.1.6.2 L-ASCORBIC ACID (CAS# 50-81-7) - Ascorbic acid reduces free chlorine at the time of sample collection (Sigma-Aldrich #255564 or equivalent).9 543-9 ------- 7.1.6.3 2-CHLOROACETAMIDE (CAS# 79-07-2) - Inhibits microbial growth and analyte degradation (Sigma-Aldrich #C0267 or equivalent).9 7.1.7 NITROGEN - Aids in aerosol generation of the ESI liquid spray and is used as collision gas in some MS/MS instruments. Nitrogen used should meet or exceed instrument manufacturer's specifications. 7.1.8 ARGON - Used as collision gas during MS/MS experiments. Argon should meet or exceed instrument manufacturer's specifications. Nitrogen gas may be used as collision gas provided sufficient sensitivity (product ion formation) is achieved. 7.2 STANDARD SOLUTIONS - When the purity of a compound is assayed to be 96% or greater, the weight can be used without correction to calculate concentration of the stock standard. The suggested concentrations are a description of concentrations used during method development, and may be modified to conform to instrument sensitivity. Standards for sample fortification generally should be prepared in the smallest volume that can be accurately measured to minimize addition of excess organic solvent to aqueous samples. Even though stability times for standard solutions are suggested in the following sections, laboratories should use standard QC practices to determine when their standards need to be replaced. 7.2.1 INTERNAL (IS) STOCK STANDARD SOLUTIONS - This method uses three IS compounds: methomyl-13C2,15N, carbofuran-13Ce, and bensulide-Ji4. These IS compounds were carefully chosen during method development because they encompass some of the functional groups of method analytes. Although alternate IS compounds may be used provided they are isotopically labeled compounds with similar functional groups as method analytes, the analyst must have documented reasons for using alternate IS compounds. Alternate IS compounds must meet the QC requirements in Section 9.3.4. 7.2.1.1 IS STOCK STANDARD SOLUTIONS (100-1000 |ig/mL) - These IS stocks can be obtained as individual certified stock standard solutions or neat materials. During development of this method, commercially obtained 100 |ig/mL stock standard solutions of methomyl-13C2,15Nin methanol (Cambridge Isotopes #CNLM-7148-1.2) and carbofuran-13Ce in 1,4-dioxane (Cambridge Isotopes # CLM-1911-1.2) were used. Bensulide- du was prepared from neat material (Cambridge Isotopes # DLM-7152) at 1000 |ig/mL in acetonitrile. IS stock standard solutions were stable for at least six months when stored at -15 °C or less in amber glass screw cap vials. 7.2.1.2 INTERNAL STANDARD PRIMARY DILUTION STANDARD (IS PDS) (32-80 pg/|iL) - Prepare, or purchase commercially, the IS PDS containing the three isotopically labeled chemicals at the suggested concentrations in the table below. If prepared from individual stock standard solutions (Sect. 7.2.1.1), the table below can be used as a guideline for preparing the IS PDS 543-10 ------- although concentrations may need to be adjusted for instrument sensitivity. The IS PDS used in these studies was prepared in acetonitrile. The IS PDS has been shown to be stable for at least six months when stored at 6 °C or less in amber glass screw cap vials. Ten jiL of this 32-80 pg/|iL IS PDS was used to fortify 10-mL samples (Sect. 11.2.2). This will yield a concentration of 32-80 ng/L of each IS in 10-mL samples. IS methomyl-13C2,15N carbofuran-13Ce bensulide-Ji4 Cone, of IS Stock (ug/mL) 100 100 1000 Vol. Of IS Stock (uL) 20 8.0 2.0 Final Vol. of IS PDS (mL) 25 Final Cone. of IS PDS (pgW 80 32 80 7.2.2 ANALYTE STANDARD SOLUTIONS - Analyte standards may be purchased commercially as ampulized solutions or prepared from neat materials. 7.2.2.1 ANALYTE STOCK STANDARD SOLUTION (1000 |ig/mL)- If preparing from neat material, accurately weigh approximately 5 mg of pure material to the nearest 0.1 mg and dilute to 5 mL with acetonitrile or methanol for a final concentration of 1000 |ig/mL. Repeat for each method analyte prepared from neat material. Alternatively, purchase commercially available individual stock standards of the analytes, preferably in methanol or acetonitrile, if available. For development of this method, all the analytes were purchased from Absolute Standards, Inc as stock standards at 1000 ug/mL. These stock standards were stable for at least 6 months when stored at -15 °C or less in amber glass screw cap vials. 7.2.2.2 INTERMEDIATE ANALYTE PRIMARY DILUTION STANDARD (PDS) SOLUTION (0.40-1.0 ng/nL) - The intermediate analyte PDS contains all, or a portion, of method analytes at various concentrations in acetonitrile. ESI and MS/MS response varies by compound; therefore, a mix of concentrations may be needed in the analyte PDS. During method development, analyte PDS solutions were prepared such that approximately the same instrument response was obtained for all analytes. The intermediate analyte PDS was prepared in acetonitrile at a concentration of 1.0 ng/|aL, except for fenamiphos, tebufenozide and tebuconazole at 0.40 ng/|aL each. The intermediate analyte PDS is prepared by dilution of the combined Analyte Stock Standard Solutions (Sect.7.2.2.1) and is used to prepare the analyte SDS (Sect 7.2.2.3). The intermediate analyte PDS has been shown to be stable for 6 months when stored at -15 °C or less in amber glass screw cap vials. 543-11 ------- 7.2.2.3 ANALYTE SECONDARY DILUTION STANDARD (SDS) (4.0-10 pg/jiL) - The analyte SDS contains all, or a portion, of method analytes at various concentrations in acetonitrile. The analyte SDS was prepared in acetonitrile at a concentration of 10 pg/i-iL, except for fenamiphos, tebufenozide and tebuconazole at 4.0 pg/|oL each. The analyte SDS is prepared by dilution of the intermediate analyte PDS (Sect.7.2.2.2) and is used to prepare CAL standards, and fortify LFSMs, LFSMDs and FDs with the method analytes. The analyte SDS has been shown to be stable for 6 months when stored at 6 °C or less in amber glass screw cap vials. 7.2.3 CALIBRATION STANDARDS (CAL) - Prepare a series of at least five concentrations of calibration solutions in deionized water, containing all the preservatives (Sect. 8.1.2), from dilutions of the analyte SDS (Sect 7.2.2.3). The suggested concentrations in this paragraph are a description of the concentrations used during method development, and may be modified to conform with instrument sensitivity. Concentrations ranging from 0.8-40 ng/L are suggested for fenamiphos, tebufenozide and tebuconazole, and 2.0-100 ng/L are suggested for the remaining analytes. Larger concentration ranges will require more calibration points. The IS is added to CAL standards at a constant concentration. During method development, the IS concentrations were 32-80 ng/L in the aqueous samples. The lowest concentration CAL standard must be at or below the MRL, which may depend on system sensitivity. CAL standards may also be used as CCCs and LFBs (Sect. 9.3.2). During method development, CAL standards were shown to be stable for four weeks when stored at 6 °C or less. Note: To avoid significant broadening of analyte peak shape, calibration standards must not contain more than 1.5% acetonitrile. 8. SAMPLE COLLECTION, PRESERVATION AND STORAGE 8.1 SAMPLE BOTTLE PREPARATION 8.1.1 Samples must be collected in amber glass bottles (100-mL or smaller) fitted with PFTE-lined screw caps. 8.1.2 Preservation reagents, listed in the table below, are added to each sample bottle as a solid prior to shipment to the field (or prior to sample collection). Compound 2-Chloroacetamide L-Ascorbic acid Trizma Amount 2.0 g/L 200 mg/L 7.75 g/L Purpose antimicrobial dechlorinating agent buffering reagent 543-12 ------- 8.2 SAMPLE COLLECTION 8.2.1 Open the cold water tap and allow the system to flush until the water temperature has stabilized (approximately 3 to 5 min). Collect samples from the flowing system. 8.2.2 Fill sample bottles, taking care not to flush out the sample preservation reagents. Samples do not need to be collected headspace free. 8.2.3 After collecting the sample, cap the bottle and agitate by hand until preservative is dissolved. Note that 2-chloroacetamide is slow to dissolve especially in cold water. Keep the sample sealed from time of collection until extraction. 8.3 SAMPLE SHIPMENT AND STORAGE - Samples must be chilled during shipment and must not exceed 10 °C during the first 48 hours after collection. Sample temperature must be confirmed to be at or below 10 °C when samples are received at the laboratory. Samples stored in the lab must be held at or below 6 °C until extraction, but should not be frozen. NOTE: Samples that are significantly above 10° C, at the time of collection, may need to be iced or refrigerated for a period of time, in order to chill them prior to shipping. This will allow them to be shipped with sufficient ice to meet the above requirements. 8.4 SAMPLE HOLDING TIMES - Water samples should be extracted as soon as possible after collection but must be extracted and analyzed within 28 days of collection. Samples must not remain at room temperature in the instrument autosampler tray for more than 29 hours prior to extraction. 9. QUALITY CONTROL 9.1 QC requirements include the Initial Demonstration of Capability (IDC) and ongoing QC requirements that must be met when preparing and analyzing Field Samples. This section describes QC parameters, their required frequencies, and performance criteria that must be met in order to meet EPA quality objectives. QC criteria discussed in the following sections are summarized in Tables 11 and 12. These QC requirements are considered the minimum acceptable QC criteria. Laboratories are encouraged to institute additional QC practices to meet their specific needs. 9.1.1 METHOD MODIFICATIONS - The analyst is permitted to modify LC columns, LC gradient, internal standards, and MS and MS/MS conditions. Each time such method modifications are made, the analyst must repeat the procedures of the IDC (Sect. 9.2). Modifications to LC conditions should still minimize co-elution of method analytes to reduce the probability of suppression/enhancement effects. Modifications to the on-line SPE cartridge, the SPE solvents and solvent volumes are not permitted. 543-13 ------- 9.1.2 If an optional on-line dual SPE cartridge system (Sect. 11.1) is used, IDC data must be obtained on both SPE cartridges and must meet the QC criteria in Sections 9.2.1-9.2.3 every time a new lot of SPE cartridges is used. NOTE: Although infrequent, recovery differences were occasionally observed, during method development, between two on-line SPE cartridges from the same lot. 9.2 INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be successfully performed prior to analyzing any Field Samples. Prior to conducting the IDC, the analyst must first generate an acceptable Initial Calibration following the procedure outlined in Section 10.2. 9.2.1 INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND - Any time a new lot of SPE cartridges, solvents, and disposable pipets are used, it must be demonstrated that an LRB is reasonably free of contamination and that criteria in Section 9.3.1 are met. 9.2.2 INITIAL DEMONSTRATION OF PRECISION (IDP) - Prepare, extract, and analyze four to seven replicate LFBs fortified near the midrange of the initial calibration curve according to the procedure described in Section 11. Sample preservatives as described in Section 8.1.2 must be added to these samples. The relative standard deviation (RSD) of the results of replicate analyses must be less than 20%. 9.2.3 INITIAL DEMONSTRATION OF ACCURACY (IDA) - Using the same set of replicate data generated for Section 9.2.2, calculate average recovery. The average recovery of replicate values must be within ± 30% of the true value. 9.2.4 MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish a target concentration for the MRL based on the intended use of the method. The MRL may be established by a laboratory for their specific purpose or may be set by a regulatory agency. Establish an Initial Calibration following the procedure outlined in Section 10.2. The lowest CAL standard used to establish Initial Calibration (as well as the low-level CCC, Section 10.3) must be at or below the concentration of the MRL. Establishing the MRL concentration too low may cause repeated failure of ongoing QC requirements. Confirm the MRL following the procedure outlined below. 9.2.4.1 Fortify, extract, and analyze seven replicate LFBs at the proposed MRL concentration. These LFBs must contain all method preservatives described in Section 8.1.2. Calculate the mean measured concentration (Mean) and standard deviation for these replicates. Determine the Half Range for the prediction interval of results (HRpm) using the equation below 543-14 ------- HRpm = 3.963^ where s = standard deviation 3.963 = a constant value for seven replicates.2 9.2.4.2 Confirm that the upper and lower limits for the Prediction Interval of Result (PIR = Mean +_ HRpm) meet the upper and lower recovery limits as shown below The Upper PIR Limit must be < 150% recovery. Mean + HRprrf ^ x 100% < 150% FortifiedConcentration The Lower PIR Limit must be > 50% recovery. Mean - HRPIR FortifiedConcentration 9.2.4.3 The MRL is validated if both the Upper and Lower PIR Limits meet the criteria described above (Sect. 9.2.4.2). If these criteria are not met, the MRL has been set too low and must be determined again at a higher concentration. 9.2.5 CALIBRATION CONFIRMATION - Analyze a QCS as described in Section 9.3.7 to confirm the accuracy of the standards/calibration curve. 9.2.6 DETECTION LIMIT DETERMINATION (optional) - While DL determination is not a specific requirement of this method, it may be required by various regulatory bodies associated with compliance monitoring. It is the responsibility of the laboratory to determine ifDL determination is required based upon the intended use of the data. Replicate analyses for this procedure should be done over at least three days. Prepare at least seven replicate LFBs at a concentration estimated to be near the DL. This concentration may be estimated by selecting a concentration at 2-5 times the noise level. DLs in Table 6 were calculated from LFBs fortified at various concentrations as indicated in the table. Appropriate fortification concentrations will be dependent upon the sensitivity of the LC/MS/MS system used. All preservation reagents listed in Section 8.1.2 must also be added to these samples. Analyze the seven replicates through all steps of Section 11. NOTE: If an MRL confirmation data set meets these requirements, a DL may be calculated from the MRL confirmation data, and no additional analyses are necessary. 543-15 ------- Calculate the DL using the following equation where s = standard deviation of replicate analyses t («-i, i-a=o.99) = Student's t value for the 99% confidence level with n-1 degrees of freedom n = number of replicates. NOTE: Do not subtract blank values when performing DL calculations. The DL is a statistical determination of precision only.3 If DL replicates are fortified at a low enough concentration, it is likely that they will not meet the precision and accuracy criteria for CCCs. Therefore, no precision and accuracy criteria are specified. 9.3 ONGOING QC REQUIREMENTS - This section summarizes ongoing QC criteria that must be followed when processing and analyzing Field Samples. 9.3.1 LABORATORY REAGENT BLANK (LRB) - An LRB is required with each extraction batch (Sect. 3.6) to confirm that potential background contaminants are not interfering with identification or quantitation of method analytes. If more than 20 Field Samples are included in a batch, analyze an LRB for every 20 samples. If the LRB produces a peak within the retention time window of any analyte that would prevent determination of that analyte, determine the source of contamination and eliminate the interference before processing samples. Background contamination must be reduced to an acceptable level before proceeding. Background from method analytes or other contaminants that inter- fere with the measurement of method analytes must be below 1/3 of the MRL. If method analytes are detected in the LRB at concentrations equal to or greater than this level, then all data for the problem analyte(s) must be considered invalid for all samples in the extraction batch. Blank contamination is estimated by extrapolation, if the concentration is below the lowest CAL standard. This extrapolation procedure is not allowed for sample results as it may not meet data quality objectives. Subtracting blank values from sample results is not permitted. 9.3.2 CONTINUING CALIBRATION CHECK (CCC) - CCC standards are analyzed at the beginning of each analysis batch, after every 10 Field Samples, and at the end of the analysis batch. See Section 10.3 for concentration requirements and acceptance criteria. 9.3.3 LABORATORY FORTIFIED BLANK (LFB) - Because this method utilizes procedural calibration standards, which are fortified reagent waters, there is no difference between the LFB and the CCC standards. Consequently, the analysis of 543-16 ------- a separate LFB is not required as part of the ongoing QC; however, the term "LFB" is used for clarity in the IDC. 9.3.4 INTERNAL STANDARDS (IS) - The analyst must monitor peak areas of the IS(s) in all injections during each analysis day. Internal standard responses (as indicated by peak areas) for any chromatographic run must not deviate by more than ± 50% from average areas measured during the initial calibration for the internal standards. If IS areas in a chromatographic run do not meet these criteria, analyze a second aliquot of that standard or sample. 9.3.4.1 If analysis of the second aliquot produces acceptable IS responses, report results for that aliquot. 9.3.4.3 If the IS area counts of the re-analyzed second aliquot still do not meet the IS criterion, the analyst should check the calibration by analyzing the most recent CCC. If the IS criterion is met in the CCC but not the sample, report the sample results as suspect/matrix. 9.3.4.4 If the IS area criterion is not met in both the sample and the CCC, instrument maintenance such as SPE cartridge replacement or sample cone cleaning may be necessary. Perform the appropriate instrument maintenance and then re- analyze the sample in a subsequent analytical batch provided the sample is still within the holding time. Otherwise, report results obtained from the re- analyzed sample, but annotate as suspect. Alternatively, collect a new sample and re-analyze. 9.3.5 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of an LFSM is required in each extraction batch and is used to determine that the sample matrix does not adversely affect method accuracy. Assessment of method precision is accomplished by analysis of a Field Duplicate (FD) (Sect. 9.3.6); however, infrequent occurrence of method analytes would hinder this assessment. If the occurrence of method analytes in samples is infrequent, or if historical trends are unavailable, a second LFSM, or LFSMD, must be prepared, extracted, and analyzed from a duplicate of the Field Sample. Extraction batches that contain LFSMDs will not require extraction of a FD. If a variety of different sample matrices are analyzed regularly, for example, drinking water from ground water and surface water sources, method performance should be established for each. Over time, LFSM data should be documented by the laboratory for all routine sample sources. 9.3.5.1 Within each extraction batch (Sect. 3.6), a minimum of one Field Sample is fortified as an LFSM for every 20 Field Samples analyzed. The LFSM is prepared by spiking a sample with an appropriate amount of the Analyte SDS (Sect. 7.2.2.3). Select a spiking concentration that is greater than or equal to the matrix background concentration, if known. Use historical data and rotate through low, mid and high concentrations when selecting a fortifying concentration. 543-17 ------- 9.3.5.2 Calculate percent recovery (%R) for each analyte using the equation c where A = measured concentration in the fortified sample B = measured concentration in the unfortified sample C = fortification concentration. 9.3.5.3 Analyte recoveries may exhibit matrix bias. For samples fortified at or above their native concentration, recoveries should range between 70-130%, except for low-level fortification near or at the MRL (within a factor of two-times the MRL concentration) where 50-150% recoveries are acceptable. If the accuracy of any analyte falls outside the designated range, and laboratory performance for that analyte is shown to be in control in CCCs, the recovery is judged to be matrix biased. The result for that analyte in the unfortified sample is labeled suspect/matrix to inform the data user that the results are suspect due to matrix effects. 9.3.6 FIELD DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (FD or LFSMD) - Within each extraction batch (not to exceed 20 Field Samples, Sect. 3.6), a minimum of one FD or LFSMD must be analyzed. Duplicates check the precision associated with sample collection, preservation, storage, and laboratory procedures. If method analytes are not routinely observed in Field Samples, an LFSMD should be analyzed rather than an FD. 9.3.6. 1 Calculate relative percent difference (RPD) for duplicate measurements (FD1 and FD2) using the equation FD1-FD2 RPD = -/ - , — xlOO FD2)/2 9.3.6.2 RPDs for FDs should be < 30%. Greater variability may be observed when the matrix is fortified at analyte concentrations at or near the MRL (within a factor of two times the MRL concentration). At these concentrations, FDs should have RPDs that are < 50%. If the RPD of any analyte falls outside the designated range, and laboratory performance for that analyte is shown to be in control in the CCC, the recovery is judged to be matrix biased. The result for that analyte in the unfortified sample is labeled suspect/matrix to inform the data user that the results are suspect due to matrix effects. 9.3.6.3 If an LFSMD is analyzed instead of a FD, calculate the relative percent difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the equation 543-18 ------- \LFSM -LFSMD\ = ^ ^ (LFSM + LFSMD)/2 9.3.6.4 RPDs for duplicate LFSMs should be < 30% for samples fortified at or above their native concentration. Greater variability may be observed when the matrix is fortified at analyte concentrations at or near the MRL (within a factor of two times the MRL concentration). LFSMs fortified at these concentrations should have RPDs that are < 50% for samples fortified at or above their native concentration. If the RPD of any analyte falls outside the designated range, and laboratory performance for that analyte is shown to be in control in the CCC, the recovery is judged to be matrix biased. The result for that analyte in the unfortified sample is labeled suspect/matrix to inform the data user that the results are suspect due to matrix effects. 9.3.7 QUALITY CONTROL SAMPLES (QCS) - As part of the IDC (Sect. 9.2), each time a new Analyte SDS (Sect. 7.2.2.3) is prepared, and at least quarterly, analyze a QCS sample from a source different from the source of the CAL standards. If a second vendor is not available, then a different lot of the standard should be used. The QCS should be prepared and analyzed just like a CCC. Fortify the QCS near the midpoint of the calibration range. Acceptance criteria for the QCS are identical to CCCs; the calculated amount for each analyte must be ± 30% of the true value. If measured analyte concentrations are not of acceptable accuracy, check the entire analytical procedure to locate and correct the problem. If the discrepancy is not resolved, one of the standard materials may be degraded or otherwise compromised and a third standard must be obtained. 10. CALIBRATION AND STANDARDIZATION 10.1 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed. After the initial calibration is successful, a CCC is required at the beginning and end of each period in which analyses are performed, and after every tenth Field Sample. 10.2 INITIAL CALIBRATION 10.2.1 ESI-MS/MS TUNE 10.2.1.1 Calibrate the mass scale of the MS with the calibration compounds and procedures prescribed by the manufacturer. 10.2.1.2 Optimize the [M+H]+ for each method analyte by infusing approximately 0.5-1.0 ng/mL of each analyte (prepared in the initial mobile phase conditions) directly into the MS at the chosen LC mobile phase flow rate (approximately 0.4 mL/min). This tune can be done on a mix of method analytes. MS parameters (voltages, temperatures, gas flows, etc.) are varied until optimal analyte responses are determined. Method analytes may have 543-19 ------- different optima requiring some compromise between the optima. See Table 3 for ESI-MS conditions used in method development. 10.2.1.3 Optimize the product ion (Sect. 3.18) for each analyte by infusing approximately 0.5-1.0 ng/mL of each analyte (prepared in the initial mobile phase conditions) directly into the MS at the chosen LC mobile phase flow rate (approximately 0.4 mL/min). This tune can be done on a mix of method analytes. MS/MS parameters (collision gas pressure, collision energy, etc.) are varied until optimal analyte responses are determined. See Table 5 for MS/MS conditions used in method development. 10.2.2 Establish on-line SPE-LC/MS/MS operating parameters as described in Tables 1, 2 and Section 11. The on-line SPE and LC conditions (solvents, volume of solvents, solvent modifiers) may not be modified. 10.2.3 Analyze a mid-level CAL standard under on-line SPE-LC/MS conditions to obtain retention times of each method analyte. Divide the chromatogram into retention time windows each of which contains one or more chromatographic peaks. During MS/MS analysis, fragment a small number of selected precursor ions ([M+H]+; Sect. 3.16) for the analytes in each window and choose the most abundant product ion. Product ions (also quantitation ions) chosen during method development are in Table 5, although these will be instrument dependent. For maximum sensitivity, small mass windows of ± 0.5 daltons around the product ion mass were used for quantitation. 10.2.4 Analyze a mid-level CAL standard under optimized on-line SPE-LC/MS/MS conditions to ensure that each method analyte is observed in its MS/MS window and that there are at least 10 scans across the peak for optimum precision. 10.2.5 Prepare a set of at least five CAL standards as described in Section 7.2.3. The lowest concentration CAL standard must be at or below the MRL, which may depend on system sensitivity. It is recommended that at least four of the CAL standards are at a concentration greater than or equal to the MRL. 10.2.6 The on-line SPE-LC/MS/MS system is calibrated using the IS technique. Use the LC/MS/MS data system software to generate a linear regression or quadratic calibration curve for each of the analytes. Curves may be concentration weighted, if necessary. 10.2.7 CALIBRATION ACCEPTANCE CRITERIA - Validate the initial calibration by calculating the concentration of each analyte as an unknown against its regression equation. For calibration levels that are < MRL, the result for each analyte should be within ± 50% of the true value. All other calibration points must calculate to be within ± 30% of their true value. If these criteria cannot be met, the analyst will have difficulty meeting ongoing QC criteria. It is recommended that corrective 543-20 ------- action is taken to reanalyze the CAL standards, restrict the range of calibration, or select an alternate method of calibration. CAUTION: When acquiring MS/MS data, on-line SPE-LC operating conditions must be carefully reproduced for each analysis to provide reproducible retention times (Sect. 11.4.1). If this is not done, the correct ions will not be monitored at appropriate times. As a precautionary measure, chromatographic peaks in each window must not elute too close to the edge of the segment time window. 10.3 CONTINUING CALIBRATION CHECK (CCC) - Minimum daily calibration verification is as follows. Verify the initial calibration at the beginning and end of each group of analyses, and after every tenth sample during analyses. In this context, a "sample" is considered to be a Field Sample. LRBs, CCCs, LFSMs, FDs and LFSMDs are not counted as samples. The beginning CCC of each analysis batch must be at or below the MRL in order to verify instrument sensitivity prior to any analyses. If standards have been prepared such that all low CAL points are not in the same CAL solution, it may be necessary to analyze two CAL standards to meet this requirement. Alternatively, analyte concentrations in the analyte SDS may be customized to meet this criteria. Subsequent CCCs should alternate between a medium and high concentration CAL standard. 10.3.1 Analyze an aliquot of the appropriate concentration CAL standard and analyze with the same conditions used during the initial calibration. 10.3.2 Determine that the absolute areas of the quantitation ions of the IS(s) are within 50-150% of the average areas measured during initial calibration. If any of the IS areas has changed by more than these amounts, adjustments must be made to restore system sensitivity. These adjustments may include cleaning of the MS ion source, or other maintenance as indicated in Section 10.3.4. Major instrument maintenance requires recalibration (Sect 10.2) and verification of sensitivity by analyzing a CCC at or below the MRL (Sect 10.3). Control charts are useful aids in documenting system sensitivity changes. 10.3.3 Calculate the concentration of each analyte in the CCC. Each analyte fortified at a level < MRL must calculate to be within ± 50% of the true value. The calculated concentration of method analytes in CCCs fortified at all other levels must be within ± 30%. If these conditions do not exist, then all data for the problem analyte must be considered invalid, and remedial action should be taken (Sect. 10.3.4) which may require recalibration. Any Field or QC Samples that have been analyzed since the last acceptable calibration verification should be reanalyzed after adequate calibration has been restored, with the following exception If the CCC fails because the calculated concentration is greater than 130% (150% for the low-level CCC) for a particular method analyte, 543-21 ------- and Field Samples show no detection for that method analyte, non-detects may be reported without re-analysis. 10.3.4 REMEDIAL ACTION - Failure to meet CCC QC performance criteria may require remedial action. Major maintenance, such as cleaning the ESI probe, atmospheric pressure ionization source, mass analyzer, replacing the LC column, replacing the SPE cartridges, etc., requires recalibration (Sect 10.2) and verification of sensitivity by analyzing a CCC at or below the MRL (Sect 10.3). CAUTION: The trizma buffer is retained to some degree on the on-line SPE cartridges and elutes over the whole chromatogram as a broad band. Thus, fouling of the ESI source cones can occur over time. It is highly recommended that the ESI probe tip be positioned as far away from the MS orifice as possible while maintaining enough sensitivity to meet the MRL. This will aid in reducing the frequency of the source cleanings. 11. PROCEDURE 11.1 This procedure applies only to fully automated on-line SPE-LC/MS/MS. Data presented in Tables 6-10 demonstrate data collected by automated on-line SPE- LC/MS/MS. This method cannot be used for manual (off-line) SPE extractions. NOTE: SPE cartridges described in this section are designed for multiple use. During method development, on-line SPE cartridges were shown to be accurate and precise up to a minimum of 330 injections. Provided QC criteria in Section 9 are met, laboratories can use the on-line SPE cartridges for longer periods. NOTE: During method development, a dual SPE cartridge system was utilized as displayed in Figures 2 and 3. The parallel two SPE cartridge system provided the capability to inject on SPE cartridge #1 while conditioning SPE cartridge #2, thereby increasing sample throughput. This dual on-line SPE cartridge system is optional. Single on-line cartridge systems may be used provided the SPE cartridge re-conditioning and re-equilibration steps in Sections 11.3.6 and 11.3.7 are followed. 11.2 SAMPLE PREPARATION 11.2.1 Samples are preserved, collected and stored as presented in Section 8. All Field and QC Samples, including the LRB, and CCC, must contain the preservatives listed in Section 8.1.2. Before extraction, verify that the sample pH is 7 ± 0.5. If the sample pH does not meet this requirement, discard the sample. If the sample pH is acceptable, proceed with the analysis. 543-22 ------- 11.2.2 Transfer a 10-mL aliquot of each sample to 20-mL (or smaller) amber autosampler vials. Add an aliquot of the IS PDS (Sect. 7.2.1.2) to each sample to be extracted, cap (caps with pre-slit septa) and invert to mix. 11.2.3 In addition to preservatives, if the sample is a CCC, FD, LFSM, or LFSMD, add the necessary amount of analyte SDS (Sect. 7.2.2.3). Cap (caps with pre-slit septa) and invert each sample to mix. 11.3 ON-LINE SPE-LC/MS/MS PROCEDURE 11.3.1 Establish operating conditions equivalent to those summarized in Tables 1-5 of Section 17. Instrument MS conditions and LC columns should be optimized prior to initiation of the IDC. 11.3.2 Inj ect a maximum of 2 mL of the preserved water sample. Inj ection volumes of less than 2 mL may be used provided sufficient sensitivity can be obtained. Smaller injection volumes will aid in reducing both matrix effects and fouling of the ESI source. During method development, a 5-mL autosampler syringe and 5-mL stainless steel loop were used to deliver the sample to the SPE cartridge. Note: During method development, the autosampler syringe was washed with 10 mL of 50:50 acetonitrile:MeOH followed by 10 mL of reagent water. Other solvents may be used, if necessary, to prevent sample carryover, but solvents may not be added to the aqueous wash as this will cause breakthrough of the analytes. 11.3.3 Load the sample in the loop onto the SPE cartridge with the loading pump using 20 mM ammonium acetate (Sect. 7.1.5) at a flow rate of 2 mL/min. With a 5-mL sample loop this takes about 3 minutes. 11.3.4 Wash the SPE cartridge with the loading pump using 20 mM ammonium acetate (Sect. 7.1.5) at a flow rate of 2 mL/min for 4.1 minutes (Table IB). 11.3.5 Elute the analytes onto the analytical column by back flushing the cartridge with the analytical pump using the gradient program in Table 1 A. The gradient steps may be modified, but not the mobile phase constituents: 20 mM ammonium acetate and acetontrile. 11.3.6 Condition the SPE cartridge (for the next injection) using the loading pump at a flow rate of 2 mL/min with acetonitrile for 2.9 min. For a dual SPE cartridge system, this conditioning step occurs concurrently (see Table IB) with the elution step in Sect. 11.3.5. For a single SPE cartridge system, the conditioning step is started after completion of the elution step in Section 11.3.5 (see Table 2). 543-23 ------- 11.3.7 Re-equilibrate the SPE cartridge to initial conditions using the loading pump at a flow rate of 2 mL/min with 20 mM ammonium acetate (Sect. 7.1.5) for 4 min. The SPE cartridge is now ready for the next injection. For a dual SPE cartridge system, this re-equilibration step occurs concurrently (see Table IB) with the elution step in Sect. 11.3.5. For a single SPE cartridge system, the re- equilibration step is started after completion of the conditioning step in Section 11.3.6 (see Table 2). NOTE: If different flow rates are used in Sections 11.3.3,11.3.4,11.3.6 and 11.3.7, then the times for each step must be modified to ensure that the same amount of solvent is used in each step. 11.4 ANALYSIS 11.4.1 Establish an appropriate retention time window for each analyte. This should be based on measurements of actual retention time variation for each method analyte in CAL standard solutions analyzed on the LC over the course of time. A value of plus or minus three times the standard deviation of the retention time obtained for each method analyte while establishing the initial calibration and completing the IDC can be used to calculate a suggested window size. However, the experience of the analyst should weigh heavily on the determination of the appropriate retention window size. 11.4.2 Calibrate the system by either the analysis of a calibration curve (Sect. 10.2) or by confirming the initial calibration is still valid by analyzing a CCC as described in Section 10.3. If establishing an initial calibration, complete the IDC as described in Section 9.2. 11.4.3 Begin analyzing Field Samples, including QC samples, at their appropriate frequency by injecting the same size aliquots (2 mL was used in method development), under the same conditions used to analyze the CAL standards. 11.4.4 At the conclusion of data acquisition, use the same software that was used in the calibration procedure to identify peaks of interest in predetermined retention time windows. Use the data system software to examine the ion abundances of the peaks in the chromatogram. Identify an analyte by comparison of its retention time with that of the corresponding method analyte peak in a reference standard. Comparison of MS/MS mass spectra is not particularly useful given the limited ±0.5 dalton mass range around a single product ion for each method analyte. 11.4.5 The analyst must not extrapolate beyond the established calibration range. If an analyte peak area exceeds the range of the initial calibration curve, the sample may be diluted with reagent water containing the preservatives. Re-analyze the diluted sample. Incorporate the dilution factor into the final concentration calculations. The resulting data should be documented as a dilution and MRLs should be adjusted accordingly. 543-24 ------- 12. DATA ANALYSIS AND CALCULATION 12.1 Complete chromatographic resolution is not necessary for accurate and precise measurements of analyte concentrations using MS/MS. In validating this method, concentrations were calculated by measuring the product ions listed in Table 5. Other ions may be selected at the discretion of the analyst. 12.2 Calculate analyte concentrations using the multipoint calibration established in Section 10.2. Do not use daily calibration verification data to quantitate analytes in samples. 12.3 Prior to reporting data, the chromatogram should be reviewed for any incorrect peak identification or poor integration. 12.4 Calculations must utilize all available digits of precision, but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty), typically two, and not more than three significant figures. NOTE: Some data in Section 17 of this method are reported with more than two significant figures. This is done to better illustrate method performance. 13. METHOD PERFORMANCE 13.1 PRECISION, ACCURACY, AND MINIMUM REPORTING LEVELS - Tables for these data are presented in Section 17. LCMRLs and DLs for each method analyte are presented in Table 6. Precision and accuracy at two concentration levels are presented for three water matrices: reagent water (Table 7); chlorinated (finished) ground water (Table 8); chloraminated/chlorinated (finished) surface water (Table 9). 13.2 SAMPLE STORAGE STABILITY STUDIES - An analyte storage stability study was conducted by fortifying the analytes into chlorinated surface water samples that were collected, preserved, and stored as described in Section 8. Precision and mean recovery (n=4) of analyses, conducted on Days 0, 7, 14, 23 and 28 are presented in Table 10. 13.3 SECOND LABORATORY DEMONSTRATION - Performance of this method was demonstrated by multiple laboratories, with results similar to those reported in Section 17. The authors wish to acknowledge the assistance of the analysts and laboratories for their participation in the multi-laboratory verification studies: 1) Don Noot and Ralph Hindle of Vogon Laboratory Services Ltd., Cochrane, AB, Canada and 2) Karen A. Randazzo, Kevin Durk, and Amanda Comando of Suffolk County Water Authority, Hauppauge, NY. 543-25 ------- 14. POLLUTION PREVENTION 14.1 This method utilizes SPE to extract analytes from water. It requires the use of very small volumes of organic solvent and very small quantities of pure analytes, thereby minimizing potential hazards to both the analyst and the environment as compared to the use of large volumes of organic solvents in conventional liquid-liquid extractions. 14.2 For information about pollution prevention that may be applicable to laboratory operations, consult "Less is Better: Guide to Minimizing Waste in Laboratories" available from the American Chemical Society's Department of Government Relations and Science Policy, 1155 16th Street N.W., Washington, D.C., 20036 or on-line at http://portal.acs.org/portal/fileFetch/CAVPCP 012290/pdf/WPCP 012290.pdf (accessed April 2014). 15. WASTE MANAGEMENT 15.1 Analytical procedures described in this method generate relatively small amounts of waste since only small amounts of reagents and solvents are used. The matrices of concern are finished drinking water or source water. However, laboratory waste management practices must be conducted consistent with all applicable rules and regulations, and that laboratories protect the air, water, and land by minimizing and controlling all releases from fume hoods and bench operations. Also, compliance is required with any sewage discharge permits and regulations, particularly the hazardous waste identification rules and land disposal restrictions. 16. REFERENCES 1. Shoemaker J.A., Tettenhorst, D.R. U.S. EPA Method 540: Determination of Selected Organic Chemicals in Drinking Water by Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS), Revision 1.0, 2013, EPA/600/R-13/119. 2. Winslow, S.D., Pepich, B.V., Martin, J.J., Hallberg, G.R., Munch, D.J., Frebis, C.P., Hedrick, E.J., Krop, R. A. "Statistical Procedures for Determination and Verification of Minimum Reporting Levels for Drinking Water Methods." Environ. Sci. Technol. 2004, 40,281-288. 3. Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, W.L. Budde, "Trace Analyses for Wastewaters" Environ. Sci. Technol. 1981, 15, 1426-1435. 4. Leenheer, J.A., Rostad, C.E., Gates, P.M., Furlong, E.T., Ferrer, I. "Molecular Resolution and Fragmentation of Fulvic Acid by Electrospray lonization/Multistage Tandem Mass Spectrometry." Anal. Chem. 2001, 73, 1461-1471. 5. Cahill, J.D., Furlong E.T., Burkhardt, M.R., Kolpin, D., Anderson, L.G. "Determination of Pharmaceutical Compounds in Surface- and Ground-Water Samples by Solid-Phase 543-26 ------- Extraction and High-Performance Liquid Chromatography Electrospray lonization Mass Spectrometry." J. Chromatogr. A, 2004, 1041, 171-180. 6. "Prudent Practices in the Laboratory: Handling and Disposal of Chemicals," National Academies Press (2011 updated version), ISBN: 9780309138642. 7. "OSHA Safety and Health Standards, General Industry," (29CFR1910), Occupational Safety and Health Administration, OSHA 2206, (Revised, July 2001). 8. "Safety in Academic Chemistry Laboratories," American Chemical Society Publication, Committee on Chemical Safety, 7th Edition. Available online at http ://portal. acs.org/portal/PublicWeb Site/about/governance/committees/chemicalsafety/ publications/WPCP 012294 (accessed April 2014). 9. Winslow, S. D. , Pepich, S. D. , Bassett, M. V., Wendelken, S. C., Munch, D. J., Sinclair, J. L. "Microbial inhibitors for U.S. EPA drinking water methods for the determination of organic compounds." Environ. Sci. TechnoL, 2001, 35, 4103-4110. 543-27 ------- 17. TABLES, DIAGRAMS, FLOWCHARTS AND VALIDATION DATA TABLE 1A. ANALYTICAL PUMP CONDITIONS (for dual cartridge system)' Time (min) Initial 3.80 4.10 9.00 9.10 Flow rate mL/min 0.40 0.010 0.40 0.40 0.40 % 20 mM Ammonium acetate 90 90 90 5.0 90 % Acetonitrile 10 10 10 95 10 Waters Acquity UPLC HSS T3 2.1 x 50 mm, 1.8 jam a The events in this table are linear gradients. The reduction in flow rate at 3.8 min is to eliminate potential pressure spikes. TABLE IB. LOADING PUMP (SPE) CONDITIONS (for dual cartridge system)" Time (min) Initial 4.10 7.00 11.00 Flow rate mL/min 2.0 2.0 2.0 2.0 % 20 mM Ammonium acetate 100 0 100 100 % Acetonitrile 0 100 0 0 a The events in this table are step movements, not gradients (Sect, 11.3.6 & 11.3.7). For example, at 4.1 minutes the pump is immediately stepped to 100% acetonitrile and is constant for 2.9 min. 543-28 ------- TABLE 2. LOADING AND SPE CONDITIONS (for single cartridge system") Loading Pumpb Time (min) Initial 4.0 4.1 9.0 9.1 12.1 12.2 16.2 Flow rate mL/min 2.0 2.0 0.10d 0.10 2.0 2.0 2.0 2.0 % 20 mM Ammonium acetate 100 100 100 100 0 0 100 100 % Acetonitrile 0 0 0 0 100 100 0 0 Analytical Pumpc Time (min) Initial 3.8e 4.1 9.0 9.1 Flow rate mL/min 0.40 0.010 0.40 0.40 0.40 % 20 mM Ammonium acetate 90 90 90 5.0 90 % Acetonitrile 10 10 10 95 10 a These single cartridge parameters were not used during method development but were extrapolated from the dual cartridge parameters for demonstration purposes. b The events in these columns are step movements, not gradients (Sect, 11.3.6 & 11.3.7). For example, at 4.1 minutes the pump is immediately stepped to 100% acetonitrile and is constant for 5 min c The events in these columns are linear gradients. dThe reduction in flowrate is optional but will minimize solvent consumption during the elution step in the single cartridge system. e The reduction in flowrate at 3.8 min is to eliminate potential pressure spikes. TABLE 3. ESI-MS METHOD CONDITIONS" ESI Conditions Polarity Capillary needle voltage Cone gas flow Nitrogen desolvation gas Desolvation gas temp. Positive ion 4kV 50 L/hr 800 L/hr 450 °C a The nomenclature used in this table is instrument specific. Other instruments may use different nomenclature. 543-29 ------- TABLE 4. METHOD ANALYTE SOURCE, RETENTION TIMES (RTs), AND SUGGESTED IS REFERENCES Analyte 3 -hydroxycarbofuran fenamiphos sulfone fenamiphos sulfoxide fenamiphos tebuconazole tebufenozide bensulide methomyl-13C2,15N (IS#1) carbofuran-13C6 (IS#2) bensulide-Ji4(IS#3) Peak# (Fig. 1) 2 3 4 6 7 8 10 1 5 9 RT (min) 5.91 6.31 6.70 7.59 7.75 8.06 8.27 5.58 6.82 8.25 IS# Ref 1 2 2 3 3 3 3 _ _ - TABLE 5. MS/MS METHOD CONDITIONS a,b Segment0 1 1 1 2 2 2 2 1 1 2 Analyte 3 -hydroxycarbofuran fenamiphos sulfoxide fenamiphos sulfone fenamiphos tebuconazole tebufenozide bensulide methomyl-13C2,15N carbofuran-13Ce bensulide-Ji4 Precursor \onA(m/z) 238 320 336 304 308 353 398 166 228 412 Product Iond'e (m/z) 181 233 266 217 70 133 314 91 171 316 Cone Voltage (v) 20 32 28 28 26 12 20 15 20 22 Collision Energyf (v) 10 26 22 24 20 22 12 10 10 12 a The nomenclature used in this table is instrument specific. Other instruments may use different nomenclature. b An LC/MS/MS chromatogram of the analytes is shown in Figure 1. c Segments are time durations in which single or multiple scan events occur. Precursor and product ions listed in this table are nominal masses. During MS and MS/MS optimization, the analyst should determine the precursor and product ion masses to one decimal place by locating the apex of the mass spectral peak place (e.g., m/z 398.1^313.9 for bensulide). These precursor and product ion masses (with one decimal place) should be used in the MS/MS method for all analyses. e Ions used for quantitation purposes. f Argon used as collision gas at a flow rate of 0.15 mL/min. 543-30 ------- TABLE 6. DLs AND LCMRLs IN REAGENT WATER Analyte 3 -hydroxycarbofuran fenamiphos sulfoxide fenamiphos sulfone fenamiphos tebuconazole tebufenozide bensulide Fortified Cone. (ng/L)a 1.0 1.0 1.0 0.4 0.4 0.4 1.0 DLb (ng/L) 0.99 0.60 0.63 0.13 0.47 0.26 0.64 LCMRLC (ng/L) 1.7 1.2 1.4 0.27 1.3 0.47 1.2 a Spiking concentration used to determine DL. Detection limits were determined by analyzing seven replicates over three days according to Section 9.2.6. c LCMRLs were calculated according to the procedure in reference 2. TABLE 7. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES IN FORTIFIED REAGENT WATER (n=5) Analyte 3 -hydroxycarbofuran fenamiphos sulfoxide fenamiphos sulfone fenamiphos tebuconazole tebufenozide bensulide methomyl-13C2,15N carbofuran-13Ce bensulide-Ji4 Fortified Cone. (ng/L) 50 50 50 20 20 20 50 80 32 80 Mean % Recovery 96.8 96.3 94.6 100 99.9 95.6 92.0 97.2 100 96.1 % RSD 9.1 1.7 5.5 4.5 3.6 5.3 2.0 5.6 2.7 4.3 Fortified Cone. (ng/L) 5.0 5.0 5.0 2.0 2.0 2.0 5.0 80 32 80 Mean % Recovery 100 102 101 105 112 101 95.2 97.6 100 94.6 % RSD 4.1 4.0 4.4 3.7 5.4 3.6 2.9 3.4 2.9 2.8 543-31 ------- TABLE 8. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES FORTIFIED FINISHED DRINKING WATER FROM A GROUND WATER SOURCE8 (n=5) Analyte 3 -hydroxycarbofuran fenamiphos sulfoxide fenamiphos sulfone fenamiphos tebuconazole tebufenozide bensulide methomyl-13C2,15N carbofuran-13Ce bensulide-Ji4 Fortified Cone. (ng/L) 50 50 50 20 20 20 50 80 32 80 Mean % Recovery 90.1 100 101 107 86.4 98.1 99.8 106 98.7 96.0 % RSD 5.5 4.0 2.2 2.7 2.6 2.5 2.0 7.1 5.1 2.9 Fortified Cone. (ng/L) 5.0 5.0 5.0 2.0 2.0 2.0 5.0 80 32 80 Mean % Recovery 98.2 106 97.6 97.9 98.1 105 100 102 99.5 94.6 % RSD 8.5 3.3 3.9 5.0 o o J.J 9.1 5.3 4.5 2.1 3.2 TOC = 0.71 mg/L and hardness = 342 mg/L as calcium carbonate TABLE 9. PRECISION AND ACCURACY DATA FOR METHOD ANALYTES FORTIFIED IN FINISHED DRINKING WATER FROM A SURFACE WATER SOURCE8 (n=5) Analyte 3 -hydroxycarbofuran fenamiphos sulfoxide fenamiphos sulfone fenamiphos tebuconazole tebufenozide bensulide methomyl-13C2,15N carbofuran-13Ce bensulide-Ji4 Fortified Cone. (ng/L) 50 50 50 20 20 20 50 80 32 80 Mean % Recovery 98.2 95.3 99.5 99.2 99.4 105 106 112 103 103 % RSD 4.7 3.2 3.8 2.1 6.9 1.1 4.7 4.3 1.9 2.2 Fortified Cone. (ng/L) 5.0 5.0 5.0 2.0 2.0 2.0 5.0 80 32 80 Mean % Recovery 108 95.4 94.1 91.9 87.6 97.5 89.8 98.3 97.4 98.3 % RSD 14 2.1 6.7 3.8 5.6 5.1 3.8 4.3 2.2 1.7 TOC = 3.20 mg/L and hardness = 68.5 mg/L as calcium carbonate 543-32 ------- TABLE 10. AQUEOUS SAMPLE HOLDING TIME DATA FOR SAMPLES OF FINISHED DRINKING WATER FROM A SURFACE WATER SOURCE3, FORTIFIED WITH METHOD ANALYTES AND PRESERVED AND STORED ACCORDING TO SECTION 8 (n=5)b Analyte 3 -hydroxycarbofuran fenamiphos sulfoxide fenamiphos sulfone fenamiphos tebuconazole tebufenozide bensulide methomyl-13C2,15N carbofuran-13Ce bensulide-Ji4 Fortified Cone. (ng/L) 50 50 50 20 20 20 50 80 32 80 DayO Mean %Rec 115 108 109 109 96.2 109 105 97.1 97.1 85.9 % RSD 3.8 3.1 3.6 1.8 4.1 4.4 2.2 2.9 2.6 2.6 Day 7 Mean %Rec 114 107 107 104 110 116 105 96.4 97.9 101 % RSD 3.8 2.6 0.9 2.9 4.7 2.1 3.8 3.0 2.2 4.7 Day 14 Mean %Rec 112 106 104 107 112 113 106 99.1 102 97.8 % RSD 5.0 2.9 3.2 2.2 4.7 0.9 3.8 2.2 1.7 1.6 Day 22 Mean %Rec 100 110 109 105 102 112 102 97.9 97.8 94.7 % RSD 7.3 4.0 4.9 6.9 5.3 4.9 8.0 5.1 5.0 5.0 Day 28 Mean %Rec 94.1 110 110 99.3 108 105 103 106 99.7 100 % RSD 3.9 2.9 3.2 3.4 2.2 3.2 3.2 3.5 1.5 1.3 a TOC = 0.72 mg/L and hardness =154 mg/L as calcium carbonate. b Internal standards were not added to samples until the day of extraction. 543-33 ------- TABLE 11. INITIAL DEMONSTRATION OF CAPABILITY QUALITY CONTROL REQUIREMENTS Method Reference Sect. 9.2.1 and 9.3.1 Sect. 9.2.2 Sect. 9.2.3 Sect. 9.2.4 Sect. 9.2.5 and 9.3.7 Sect. 9.2.6 Requirement Initial Demonstration of Low System Background Initial Demonstration of Precision (IDP) Initial Demonstration of Accuracy (IDA) Minimum Reporting Limit (MRL) Confirmation Quality Control Sample (QCS) Detection Limit (DL) Determination (optional) Specification and Frequency Analyze LRB prior to any other IDC steps. Analyze four to seven replicate LFBs fortified near the midrange calibration concentration. Calculate average recovery for replicates used in IDP. Fortify, extract and analyze seven replicate LFBs at the proposed MRL concentration. Calculate the Mean and the Half Range (HR). Confirm that the upper and lower limits for the Prediction Interval of Result (Upper PIR, and Lower PIR, Sect. 9.2.4.2) meet the recovery criteria. Analyze a standard from a second source, as part of IDC. Over a period of three days, prepare a minimum of seven replicate LFBs fortified at a concentration estimated to be near the DL. Analyze the replicates through all steps of the analysis. Calculate the DL using the equation in Sect. 9.2.6. Acceptance Criteria Demonstrate that all method analytes are below 1/3 the MRL and that possible interferences from extraction media do not prevent the identification and quantification of method analytes. %RSDmustbe <20% Mean recovery + 30% of true value Upper PIR < 150% Lower PIR > 50% Results must be within 70-130% of true value. Data from DL replicates are not required to meet method precision and accuracy criteria. If the DL replicates are fortified at a low enough concentration, it is likely that they will not meet precision and accuracy criteria. NOTE: Table 11 is intended as an abbreviated summary of QC requirements provided as a convenience to the method user. Because the information has been abbreviated to fit the table format, there may be issues that need additional clarification, or areas where important additional information from the method text is needed. In all cases, the full text of the QC in Section 9 supersedes any missing or conflicting information in this table. 543-34 ------- TABLE 12. ONGOING QUALITY CONTROL REQUIREMENTS (SUMMARY) Method Reference Sect. 8.1- Sect. 8.4 Sect. 9.3.1 Sect. 9.3.4 Sect. 9.3.5 Sect. 9.3.6 Sect. 9.3.7 Sect. 10.2 Sect. 9.3.2 and Sect. 10.3 Requirement Sample Holding Time Laboratory Reagent Blank (LRB) Internal Standard (IS) Laboratory Fortified Sample Matrix (LFSM) Laboratory Fortified Sample Matrix Duplicate (LFSMD) or Field Duplicates (FD) Quality Control Sample (QCS) Initial Calibration Continuing Calibration Check (CCC) Specification and Frequency 28 days with appropriate preservation and storage as described in Sections 8.1-8.4. Analyze one LRB with each extraction batch of up to 20 field samples Internal standards, methomyl-13C2,15N, carbofuran- 13Ce, andbensulide-<5?i4, are added to all standards and sample extracts, including QC samples. Compare IS areas to the average IS area in the initial calibration. Analyze one LFSM per extraction batch (20 samples or less) fortified with method analytes at a concentration greater than or equal to the native concentration, if known. Calculate LFSM recoveries. Extract and analyze at least one FD or LFSMD with each extraction batch (20 samples or less). A LFSMD may be substituted for a FD when the frequency of detects are low. Calculate RPDs. Analyze at least quarterly or when preparing new standards, as well as during the IDC. Use IS calibration technique to generate a first or second order calibration curve. Use at least five standard concentrations. Check the calibration curve as described in Sect. 10.2.7. Verify initial calibration by analyzing a low level (at the MRL or below) CCC prior to analyzing samples. CCCs are then injected after every 10 Field Samples and after the last sample, rotating concentrations to cover the calibrated range of the instrument. Acceptance Criteria Sample results are valid only if samples are extracted within the sample holding time. Demonstrate that all method analytes are below 1/3 the MRL, and confirm that possible interferences do not prevent quantification of method analytes. If targets exceed 1/3 the MRL or if interferences are present, results for these subject analytes in the extraction batch are invalid. Peak area counts for all ISs in all injections must be within + 50% of the average peak area calculated during the initial calibration. If ISs do not meet this criterion, corresponding target results are invalid. For LFSMs fortified at concentrations < 2 x MRL, the calculated recovery must be within ± 50% of the true value. At concentrations greater than the 2 x MRL, the recovery must be ± 30% of the true value. If these criteria are not met, results are labeled suspect due to matrix effects. For LFSMDs or FDs, the calculated relative percent difference must be < 30%. (< 50% if concentration < 2 x MRL.) If these criteria are not met, results are labeled suspect due to matrix effects. Results must be within 70-130% of true value. When each CAL standard is calculated as an unknown using the calibration curve, the analyte results must be 70-130% of the true value for all except CAL standards < MRL, which must be 50-150% of the true value. It is recommended that corrective action is taken to reanalyze the CAL standards, restrict the range of calibration, or select an alternate method of calibration. Each analyte fortified at a level < MRL must calculate to be within ± 50% of the true value. The calculated concentration of the method analytes in CCCs fortified at all other levels must be within ± 30%. NOTE: Table 12 is intended as an abbreviated summary of QC requirements provided as a convenience to the method user. Because the information has been abbreviated to fit the table format, there may be issues that need additional clarification, or areas where important additional information from the method text is needed. In all cases, the full text of the QC in Section 8-10 supersedes any missing or conflicting information in this table. 543-35 ------- FIGURE 1. EXAMPLE CHROMATOGRAM (OVERLAID MS/MS SEGMENTS) OF A CALIBRATION STANDARD WITH METHOD 543 ANALYTES AT CONCENTRATION LEVELS OF 3.2-8.0 ug/L. NUMBERED PEAKS ARE IDENTIFIED IN TABLE 4. 100 9,10 u C 93 •a e = 4.00 4.50 5.00 5.50 6,00 6.50 7.00 Retention Time, min 7.50 8,00 8.50 9.00 543-36 ------- FIGURE 2. ON-LINE SPE-LC/MS/MS DIAGRAMS FOR A DUAL CARTRIDGE SPE SYSTEM SPE Injection port Analytical Injection port ToLC Loading pump SPE Injection port Analytical pump Analytical Injection port ToLC Optional To waste Loading pump Analytical pump B SPE Injection port Analytical Injection port ToLC Loading pump Analytical pump A: Sample injection B: Sample loading and washing of SPE cartridge C: Sample elution while conditioning and equilibrating optional second SPE cartridge Diagrams courtesy of Waters Corp. 543-37 ------- FIGURE 3. ON-LINE SPE EVENTS FOR AN SPE DUAL CARTRIDGE SYSTEM Start Sample loading 20 mM NH4OAc Loading Pump SPE wash 20 mM NH4OAc Loading Pump 5 min gradient elution 20 mM NH4OAc ACN Re-equilibration Analytical Pump SPE Cartridge #1 Analytical Pump 10 nTime, min 12 Start > 1 ' 1 Analytical column initial conditions . 20 mM NH.OAc 4 ACN Analytical Pump K ^ 9 2 Re-condition J ACN Loading Pump f \ 4 6 Re-equilibration J 20 mM NH4OAc Loading Pump f > 8 10 SPE Cartridge #2 f Timp m 12 543-38 ------- |