EPA/625/R-96/010a
      Compendium of Methods
       for the Determination of
        Inorganic Compounds
           in Ambient Air
            Chapter IO-5

  SAMPLING AND ANALYSIS
FOR ATMOSPHERIC MERCURY
      Center for Environmental Research Information
         Office of Research and Development
         U.S. Environmental Protection Agency
             Cincinnati, OH 45268

               June 1999

-------
[This page intentionally left blank]

-------
                             EPA/625/R-96/010a
       Compendium of Methods
       for the Determination of
        Inorganic Compounds
           in Ambient Air
      Compendium Method IO-5

    SAMPLING AND ANALYSIS
    FOR VAPOR AND PARTICLE
PHASE MERCURY IN AMBIENT AIR
 UTILIZING COLD VAPOR ATOMIC
 FLUORESCENCE SPECTROMETRY
             (CVAFS)
       Center for Environmental Research Information
         Office of Research and Development
         U.S. Environmental Protection Agency
            Cincinnati, OH 45268

              June 1999

-------
                                       Method 10-5

                                    Acknowledgments

This Method is part of the Compendium of Methods for the Determination of Inorganic Compounds in
Ambient Air (EPA/625/R-96/010a), which was prepared under ContractNo. 68-C3-0315, WANo. 2-10, by
Midwest Research Institute (MRI),  as a subcontractor to Eastern Research Group, Inc. (ERG), and under the
sponsorship of the U.S. Environmental Protection Agency (EPA). Justice A. Manning, John 0. Burckle,
Scott Hedges, Center for Environmental Research Information (CERI), and Frank F. McElroy, National
Exposure Research  Laboratory (NERL), all  in the EPA's Office of Research and Development, were
responsible for  overseeing the preparation of  this method. Other support was provided by the following
members of the Compendia Workgroup:

    •    James L. Cheney, U.S. Army Corps of Engineers, Omaha, NE
        Michael F. Davis, U.S. EPA, Region 7, KC, KS
    •    Joseph B. Elkins Jr., U.S. EPA, OAQPS, RTF, NC
    •    Robert G. Lewis, U.S. EPA, NERL, RTF, NC
    •    Justice A. Manning, U.S. EPA, ORD, Cincinnati, OH
    •    William A. McClenny, U.S. EPA, NERL, RTF, NC
    •    Frank F. McElroy, U.S.  EPA, NERL, RTF, NC
    •    William T. "Jerry" Winberry, Jr., EnviroTech Solutions, Gary, NC

This Method is the result of the efforts of many individuals.  Gratitude goes to each person involved in the
preparation and review of this methodology.
Author (s)
        Gerald Keeler, University of Michigan, Ann Arbor, MI
        Jim Barres, University of Michigan, Ann Arbor, MI
Peer Reviewers
        Susan Kilmer, Michigan Department of Natural Resources, Lansing, MI
        Eric Prestbo, Frontier GeoSciences, Seattle, WA
        Anne M. Falke, Frontier GeoSciences, Seattle, WA
        Jamie Brown, Supelco Inc., Bellefonte, PA
        Alan Zaffiro, International Technology Corporation,  Cincinnati, OH
        Dr. Matt Landis, U.S. EPA, NERL, RTF, NC
        Lauren Drees, U.S. EPA, NRMRL, Cincinnati, OH
                                       DISCLAIMER
This Compendium has been subjected to the Agency's peer and administrative review, and it has
been approved for publication as an EPA document. Mention of trade names or commercial products
does not constitute endorsement or recommendation for use.
                                            11

-------
                                       Method IO-5
             Sampling and Analysis for Vapor and Particle Phase Mercury in
                 Ambient Air Utilizing Cold Vapor Atomic Fluorescence
                                  Spectrometry (CVAFS)

                                 TABLE OF CONTENTS
1.  Scope    	     5.0-1
2.  Applicable Documents	     5.0-2
     2.1  ASTM Documents	     5.0-2
     2.2  Other Documents	     5.0-2
3.  Method Summary	     5.0-2
4.  Significance	     5.0-3
5.  Definitions  	     5.0-4
6.  Contamination	     5.0-6
7.  Interferences	     5.0-7
8.  Safety, Restrictions, and Limitations  	     5.0-7
9.  Facilities, Equipment, and Materials  	     5.0-8
     9.1  Facilities  	     5.0-8
     9.2  Equipment	     5.0-8
     9.3  Materials	     5.0-9
10. Preparation of Supplies, Adsorbents and Reagents	    5.0-11
     10.1 Acid Cleaning Procedure	    5.0-11
     10.2 Preparation of Gold-Coated Beads and Trap  	    5.0-11
     10.3 Preparation of Glass Fiber Filters	    5.0-12
11. Collection of Ambient Air Samples  	    5.0-13
     11.1 The Sampling Equipment	    5.0-13
     11.2 Collection of Vapor Phase Hg Samples	    5.0-14
     11.3 Collection of Particulate Hg Samples	    5.0-16
     11.4 Sample Storage	    5.0-18
12. Analysis of Ambient Air Samples	    5.0-19
     12.1 Introduction  	    5.0-19
     12.2 The Analytical System  	    5.0-19
     12.3 Preparation of Reagents and Standards	    5.0-20
     12.4 Summary of Dual-amalgamation CVAFS Analytical Procedure  	    5.0-21
     12.5 Analysis of Vapor Phase Hg Samples	    5.0-21
     12.6 Analysis of Hg in Particulate Samples  	    5.0-22
13. Calculation of Mercury Concentrations in Ambient Air  	    5.0-25
     13.1 Calculation of Vapor Phase Mercury Concentrations	    5.0-25
     13.2 Calculation of Particle-Phase Mercury Concentration  	    5.0-27
14. Quality Assurance/Quality Control (QA/QC)	    5.0-28
     14.1 Personnel Qualifications 	    5.0-28
     14.2 QA/QC Samples  	    5.0-28
     14.3 Precision and Accuracy	    5.0-29
                                             ill

-------
                      TABLE OF CONTENTS (continued)
15. References	    5.0-30
                                    IV

-------
                                         Chapter 10-5
                 SAMPLING AND ANALYSIS FOR ATMOSPHERIC MERCURY

                                         Method 10-5
        SAMPLING AND ANALYSIS FOR VAPOR AND PARTICLE PHASE MERCURY IN
              AMBIENT AIR UTILIZING COLD VAPOR ATOMIC FLUORESCENCE
                                      SPECTROMETRY
1. Scope

1.1 Elemental mercury (Hg) and most of its derivatives are metabolic poisons which bioaccumulate in aquatic
food chains, ultimately reaching concentrations capable of causing neurological and reproductive damage
in terrestrial, as well as, aquatic organisms.  Atmospheric Hg, although present only in trace amounts, has
been established as a significant source of Hg to aquatic environments.

1.2 The widespread use of Hg stems largely from its electrical conductivity, high specific gravity, and
fungicidal properties. The major sources of atmospheric Hg include combustion processes (incineration of
medical waste, municipal waste, sewage sludge, and hazardous waste, as well as burning of fossil fuels), and
manufacturing processes (iron and steel production, mining/smelting operations, cement production, and coke
production).

1.3 Mercury compounds in the atmosphere exist in vapor and particulate forms, preferentially partitioning
into the vapor phase.  Mercury species fall within two main categories: inorganic Hg compounds and organic
Hg compounds.  The most common form of inorganic Hg is elemental Hg vapor. Other inorganic Hg forms
include mercuric chloride (HgCl2) and mercurous chloride (HgCl). The organic compounds include those
compounds in which Hg is covalently bonded to a carbon atom, as in the case of methyl and dimethyl Hg.

1.4 Increased focus on atmospheric Hg pollution has resulted from the Clean Air Act Amendments of 1990.
Mercury and its compounds are included in the Title III list of hazardous  air pollutants and are subject to
standards established under Section 112, including maximum achievable control technology (MACT). Also,
Section  112(c)(6) of the 1990 Amendments mandates that Hg (among others) be subject to standards that
allow for the maximum degree of emission reduction. These standards are to be promulgated no later than
10 years following the date of enactment. Additionally, within 5 years of the date of enactment, a list of
source categories that account for no less than 90 percent of Hg emissions  must be established.

1.5 As a result of the congressional mandates, the U.S. Environmental Protection Agency (EPA), state, and
local air pollution control agencies are under increased pressure to monitor the trace levels of mercury in the
ambient air.  In addition, attempts to  identify source/receptor relationships for these  substances and to
develop control programs to regulate emissions have been initiated.

1.6 Previous methods used to collect vapor-phase Hg have relied on the rapid amalgamation between Hg and
gold or silver and in some cases the adsorption of Hg onto activated charcoal. Mercury was then quantified
using atomic absorption spectrometry or instrumental neutron activation analysis.  These methods generally
required long-duration samples due to higher detection limits. Collection of vapor-phase Hg was improved
with the use of thin gold plating  on sand packed in a trap to maximize the  surface area  for collection
(Fitzgerald  and Gill,  1979).  Improvements to analytical detection limits were also made by employing cold-
vapor atomic fluorescence spectrometry (Bloom and Fitzgerald, 1988).
June 1999              Compendium of Methods for Inorganic Air Pollutants             Page 5.0-1

-------
Method IO-5.0                                                                     Chapter IO-5
Vapor and Particulate Mercury	Mercury

1.7 Previous methods used to collect particulate Hg included trapping particles on quartz wool, which was
susceptible to artifact formation due to the adsorption of vapor-phase Hg.  Quartz-wool plugs should not be
utilized for ambient particulate Hg sampling.  Additional methods have included air filtration through quartz
fiber filters followed by acid extraction (see Inorganic Compendium Methods 10-2 and 10-3). The acid
extraction techniques have,  until recently, been plagued by high Hg concentrations in the acid extraction
solutions.  Acid extraction can be utilized as  a suitable alternative to the microwave digestion technique in
many applications where precision requirements and detection limits are not limiting factors. Instrumental
neutron activation analysis has also been used to quantify mercury in particulates collected on Teflon® filters
(Lamborg et  al.,  1994).  Microwave  digestion  (see Inorganic Compendium Method  10-3.1) has been
demonstrated to give equivalent results to acid extraction.

1.8 This method describes procedures for collection and analysis of vapor phase and particulate Hg,  in order
to provide an EPA-approved, accessible sampling and analytical methodology, for uniform monitoring of
atmospheric mercury levels. These procedures have been used successfully in long-term regional monitoring
programs, as well as short-term intensive studies. For best results, it is suggested that Method 10-5 be used
over Methods 10-2 or 10-3 for sampling and analysis of particulate-phase Hg.
2. Applicable Documents

2.1 ASTM Documents

   •   D1356 Definition of Terms Related to Atmospheric Sampling and Analysis
   •   D1357 Practice for Planning the Sampling of the Ambient Atmosphere

2.2 Other Documents

   •   U. S. Environmental Protection Agency, Quality Assurance Handbook for Air Pollution Measurement
      Systems, EPA-600/9-76-005.
   •   U. S. Environmental Protection Agency, Quality Assurance Handbook for Air Pollution Measurement
      Systems, EPA-600/4-77-027a.
   •   Scientific Publications of Ambient Air Studies (1-12).
3. Method Summary
[Note:   This method  for collection and analysis of mercury in  ambient air takes  advantage of the
amalgamating property of mercury to a gold surface.  In the following description reference is made to gold-
coated glass beads only.  However, other media have been shown to give equivalent results.  These include,
but are not limited to, gold-coated sand and a solid gold matrix.]
Page 5.0-2             Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury	Vapor and Particulate Mercury

3.1 The collection of mercury, from ambient air in the vapor and particulate phase, involves use of gold-
coated bead traps and glass-fiber filters. The amalgamation process for vapor-phase mercury requires a flow
rate low enough to allow adsorption of the mercury in the air to the gold surface.  The significantly lower
levels of particle-phase mercury on the other hand require  a much higher flow rate in order  to collect
sufficient particle mass for mercury determination. Therefore, separate sampling systems are needed for the
collection of mercury in the vapor and particle phases. Accurate flow determinations through both sampling
systems are critical in providing accurate Hg concentrations in air.

3.2 Vapor-phase Hg is collected using gold-coated glass bead traps.  A Teflon® filter pack with a glass fiber
filter is  placed in front of the traps to remove particulate material from the air being sampled.  Air is pulled
through the vapor-phase sampling system using a mass-flow controlled vacuum pump at a nominal flow rate
of 0.3 Lpm.

3.3 Particle-phase Hg is collected using a glass-fiber filter in an open-faced Teflon® filter pack. Air is pulled
through the particulate sampling system using a vacuum pump at a nominal flow rate of 30 Lpm.

3.4 Determination of vapor- and particle-phase mercury  in ambient air is accomplished using cold-vapor
atomic fluorescence spectrometry (CVAFS); more specifically, dual-amalgamation CVAFS.  The amount
of vapor-phase mercury collected on a gold-coated bead trap is determined directly by CVAFS. The sample
trap is heated to release the collected mercury.  The desorbed mercury is carried in an inert gas stream (He
or Ar) to a second gold-coated bead trap, the analytical trap.  The mercury collected on the analytical trap
is  then thermally desorbed and carried into the CVAFS analyzer.  The resulting voltage peak is integrated
to  produce peak area for the sample.

3.5 Determination of Hg in the particle phase requires acid extraction of the glass-fiber filters prior to
analysis. The sample filters are extracted in a nitric acid solution using microwave digestion to yield "acid-
extractable" particulate mercury.  The extract is oxidized with BrCl to convert all forms of Hg to Hg+2 and
SnCl2 is added to the extract to  reduce the Hg+2 to volatile Hg°. The Hg° is liberated from the extract by
purging with  an inert gas (N2) and collected on a gold-coated bead trap.  The amount of mercury collected
on the trap is then  determined using dual-amalgamation CVAFS.   The detection  limits achieved using
Method 10-5  are 30 pg/m3 for particulate Hg and 45 pg/m3 for vapor mercury.
4. Significance

4.1 The area of toxic air pollutants has been the subject of interest and concern for many years.  Recently
the use of receptor models has resolved the elemental composition of atmospheric aerosol into components
related to emission sources.  The assessment of human health impacts resulting in major decisions on control
actions by federal, state, and local governments are based on these data.

4.2 Elemental mercury and most compounds of mercury are protoplasmic poisons and, therefore,  may be
lethal to all forms of living matter.  In general, the organic mercury compounds are more toxic than mercury
vapor or the inorganic compounds. Even small amounts of mercury vapor or many mercury compounds can
produce mercury intoxication when inhaled by man.  Acute mercury poisoning, which can be fatal or cause
permanent damage to the nervous system, has resulted from inhalation of from 1,200 to 8,500 Mg/m3 of
mercury. The more common chronic poisoning (mercurialism) which also affects the nervous system is an
June 1999             Compendium of Methods for Inorganic Air Pollutants             Page 5.0-3

-------
Method IO-5.0                                                                     Chapter IO-5
Vapor and Particulate Mercury	Mercury

insidious form in which the patient may exhibit no well-defined symptoms for months or sometimes years
after exposure.  The symptoms usually associated with mercurialism are exaggerated emotional response,
gingivitis, and muscular tremors.  A person suffering from a mild case of mercury poisoning is usually
unaware of the cause of the illness because the symptoms are psychopathological in nature.  Likewise, these
ambiguous symptoms may result in an incorrect diagnosis by a physician.  In addition, animals and plants
also exhibit a low tolerance to mercury and its compounds.
5. Definitions

5.1 Calibration-the process of correlating instrument response to known standard units of measure by
regression analysis for the purpose of quantifying unknown samples based on observed instrument readings.

5.2 Calibration Standard-a solution prepared from a certified mercury standard stock solution which is
used to calibrate instrument response to concentration.

5.3 Laboratory Duplicates  (LD1 and LD2)—two aliquots of the same sample treated exactly the same
throughout preparation and analysis. Analyses of laboratory duplicates provide an indication of precision
associated with  laboratory procedures.

5.4 Collocated  Samples—samples collected from identical sampling systems placed in the field side-by-side.
Collocated samples are collected and analyzed identically and are used to determine overall sampling and
analytical precision.

5.5 Reagent Blank—the reagent blank is used to determine the amount of mercury added to samples from
the laboratory reagents.

5.6 Field Blank (FB)—a field blank is treated exactly like an ambient  sample during sample setup and
removal, but air is not drawn through the field blank. The FB is used to determine the limit of detection and
is site-specific.

5.7 Storage Blank (STB)—a blank created by storing either a gold-coated trap (vapor phase Hg), or a filter
(particle phase Hg) prior  to analysis. The STB is used to determine the amount of mercury added to samples
during storage.

5.8 Mercury Standard  Stock—a solution containing a certified concentration of mercury  obtained from a
commercial source (e.g., NIST-3133).  This standard is used to prepare calibration standards.

5.9 Precision-a measure of the overall uncertainty in a particular measurement.  Assessment of precision
requires the analysis of collocated samples.
Page 5.0-4             Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury _ Vapor and Particulate Mercury

Precision is assessed using relative percent difference (RPD) between the collocated samples.  The average
precision is reported as the average of the absolute value of the RPD.
                                              (X1+X2)/2

where:

   Xj = concentration from the first collector; and
   X2 = concentration from the second collector.

5.10 Accuracy-a measure of the degree to which a measurement or computed value reflects the true value
of analyte present. Accuracy is assessed using standard reference materials (SRM) that have been processed
in a manner identical to the field samples.

Percent recovery, R, is used to assess accuracy from standard reference materials.  Recovery is calculated
as:

   R = (Measured Mass/Actual Mass) (100)

Certified reference materials for mercury in vapor and particulate phases are not available, preventing the
assessment of accuracy in these media.  Interlaboratory calibrations and sample exchanges are strongly
recommended.

5.11 Detectability— the detection limit is defined as the lowest value of a characteristic that a measurement
process, or a method-specific  procedure  can reliably discern. Three types of detection limits include the
instrument detection limit (IDL) , the method detection limit (MDL) ,  and the system detection limit (SDL)
defined as:
   5.11.1  IDL.  The smallest signal above background noise that an instrument can detect at a 99 percent
confidence level. This value is quantified by direct injection of a spiked blank or other low level sample (1-5
times the theoretical IDL) into  the instrument.  The sample does not undergo any sample processing steps of
the analytical methods and is used to provide the detection capabilities of the instrument.
   5.11.2  MDL.  The minimum concentration of a substance measured  and reported with 99 percent
confidence that the analyte concentration is greater than zero.  This is determined from the analysis of a
sample in  a given matrix containing the analyte.  It is essential that all sample  processing steps of the
analytical method be included in the determination of the MDL.  In many cases the term "limit of detection"
(LOD) and MDL are used synonymously.
   5.11.3  SDL.  The  minimum concentration of analyte  measured in a  sample that is detectable and
distinguishable from background noise of the entire  data collection system.  For Hg analyses, the SDL is
defined as either the mean or 3o of the field blanks, whichever value is largest.

5.12 Completeness-The measure of the number of valid samples  (meeting all QA requirements) obtained
compared  to the number required to achieve the objectives of the study. Overall completeness is the number
of valid samples compared to the number planned.  Laboratory completeness is the number of valid samples
obtained compared to the number analyzed. As with the other date quality attributes, completeness can be
controlled through adherence to analysis protocols in order to minimize contamination and sampling errors.
June 1999              Compendium of Methods for Inorganic Air Pollutants             Page 5.0-5

-------
Method IO-5.0                                                                      Chapter IO-5
Vapor and Particulate Mercury	Mercury

Completeness is calculated as:

                                    Completeness = v/(n)(100)

where:

   v = number of samples judged valid
   n = total number of measurements necessary to achieve project objectives


6. Contamination

6.1 Determination of Hg in atmospheric samples requires the ability to reliably detect very small (picogram)
amounts of mercury. The potential for contamination of ambient samples cannot be overemphasized.

6.2 Much of the  effort required to obtain good data is associated with maintaining the cleanest possible
conditions. Absolute adherence to clean  techniques is essential throughout all phases of sample collection,
handling,  and analysis.

   6.2.1 Class 100 Clean Room-A key element in minimizing exposure of samples to potential contaminants
includes the use of a metal-free Class  100 clean room, which is supplied by HEPA filtered air that has been
passed through charcoal adsorbent to reduce vapor-phase Hg levels.  Procedures conducted in the clean room
include the final stages of acid-cleaning for sample containers and supplies, the  extraction and analysis of
particulate mercury  samples, and storage of ultra-pure reagents.
   6.2.2  Preparation of Sampling Supplies to Minimize Contamination-All containers with which the
sample comes into contact are acid-cleaned using the protocol described in this method.  Supplies used to
manufacture gold-coated traps for vapor-phase Hg collection and filters used for collection of particulate Hg
are baked prior to use to volatilize Hg.   Particle-free gloves are worn during all manipulations of sample
containers and supplies.   All acid-cleaned supplies  are  stored double  or triple-bagged in  resealable
polyethylene. Containers for gold-coated bead traps and glass fiber filters prepared for sampling are sealed
with Teflon® tape and-bagged.
   6.2.3  Minimizing Exposure During Sample Collection-On-site preparation of filters and traps for
sampling is conducted by trained operators, outdoors to minimize potential particulate contributions. Sample
preparation can also  be performed indoors if a Class 100 environment is utilized. During all phases of sample
set-up and removal,  the operator wears particle-free gloves and  stands downwind of the sample in order to
prevent contamination by shedding particles from clothing or breathing on samples. After sampling,  filters
and traps are re-sealed in containers using Teflon® tape and bagged in resealable polyethylene.
   6.2.4  Minimizing Exposure During Sample Analysis-All reagents utilized in the analysis of samples for
Hg are highly purified to minimize Hg  content.  Suggested sources for low-Hg containing reagents are
supplied  in this method.  Blank levels  in  reagents are carefully monitored and reagents are replaced
periodically to maintain the lowest levels possible.  Particle-free gloves are worn during all phases of sample
analysis.  Blanks from extraction vessels for particulate Hg analysis are monitored and analytical systems are
maintained to insure consistent, low detection limits.
Page 5.0-6              Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury	Vapor and Particulate Mercury
7. Interferences

7.1  Types of interferences and the procedures used to resolve them analytically vary depending on the
manufacturer and model of the analytical instrument used. There are no known positive interferences at the
wavelength of 253.7 nm used to  excite and measure the fluorescence of Hg atoms using the CVAFS
procedure as described.  Suspected negative interferences include polyaromatic hydrocarbons and water
vapor. Excessive water vapor will interfere by quenching the fluorescence signal as the Hg is liberated from
the gold trap.  Free halogens also present a hazard to the gold traps resulting in a permanent destruction of
the trap resulting in low values or  no response at all.

7.2 Ambient levels of dimethyl mercury,  sulfur dioxide, hydrogen sulfide, and nitrogen dioxide do not
interfere  with the collection or analysis of mercury vapor when utilizing silver wool techniques.  These
compounds have not, however, been thoroughly tested utilizing the gold-coated bead trap technique. It is the
users  responsibility to evaluate  these compounds when using this methodology.
8. Safety, Restrictions, and Limitations

8.1 The toxicity or carcinogenicity of reagents used in this method have not been fully established.  Each
chemical should be regarded as a potential health hazard and exposure to these compounds should be as low
as reasonably achievable. Each laboratory is responsible for maintaining a current awareness file of OSHA
regulations regarding the safe handling of the chemicals specified in this method. A reference file of material
data handling sheets should also be available to all personnel involved in the chemical analysis.

8.2 Mercury compounds are highly toxic if inhaled, swallowed or absorbed through the skin.  Laboratory
personnel  should use caution and wear gloves when handling standards containing mercury.

8.3 The American Conference of  Governmental Industrial Hygienists (ACGIH) has adopted the threshold
limit value (TLV) of 100 p.g/m3 for mercury vapor and inorganic compounds of mercury for an 8-hour work
day. In addition, the ACGIH has established a TLV for organic mercury of 10 p-g/m3 for an 8-hour exposure.
9. Facilities, Equipment, and Materials

9.1 Facilities

   9.1.1 Clean room-Class 100 with down-flow, positive pressure ventilation, and separated dressing room.
Construction materials must be nonmetallic, preferably plastic or varnished wood attached without metal
fasteners.   Metal parts for which no substitute exists must be painted or otherwise sealed.  Paints and
varnishes must not contain mercury fungicides or additives. Non-permanent plastic enclosures which provide
HEP A-filtered air at adequate flow rates can also, under many circumstances, provide an adequate clean
environment.  These  facilities will likely require more strict adherence to maintenance  and  cleaning
schedules.
June 1999             Compendium of Methods for Inorganic Air Pollutants             Page 5.0-7

-------
Method IO-5.0                                                                     Chapter IO-5
Vapor and Particulate Mercury	Mercury

      9.1.1.1 Adhesive mats, for use at entry points and at work stations to remove dust and dirt from clean-
room boots.
      9.1.1.2 Laminar-flow exhaust hood for clean reagent preparation and particulate Hg sample processing
and purging.
      9.1.1.3  Clean room suit, hood, boots, and particle free gloves.
   9.1.2  Fume hoods - for the acetone and hydrochloric  acid soaking steps of the labware cleaning system.

9.2 Equipment

[Note: Following is a list of the required facilities, equipment, supplies and reagents for sample preparation,
collection, and analysis.   The make and model of some of the items are provided although many of the
materials are available from a variety of sources.]

   9.2.1  Mercury Analysis System

[Note: Separate CVAFS analysis systems dedicated to measurement of mercury from vapor phase  and
particulate samples is recommended since standardization procedures and analytical levels ofHg differ for
the two media.]

      9.2.1.1  Cold vapor atomic fluorescence spectrometer (CVAFS).
      9.2.1.2  Power conditioner to reduce voltage fluctuations to CVAFS.
      9.2.1.3  Integrator.
      9.2.1.4  Variable voltage transformers (2) for thermally desorbing traps.
      9.2.1.5  Mass flow controller for maintaining constant flow rate of carrier gas through CVAFS.
      9.2.1.6  Four-channel programmable circuit controller.
      9.2.1.7  Axial fans  (2), 30 CFM, for cooling traps.
      9.2.1.8  Nickel-chromium wire coils (2) for thermally desorbing gold traps.
   9.2.2  Microwave Digestion System
      9.2.2.1  Microwave oven equipped with temperature and pressure control system for extracting
particulate Hg samples.
      9.2.2.2  Teflon® digestion vessels.
   9.2.3  Muffle furnace-for baking glass fiber filters,  quartz wool, and quartz tubes; capable of heating to
700°C.
   9.2.4  Pumps and Flow Measurement System
      9.2.4.1  Sampling box (plastic enclosure, suitable for outdoor conditions,  approximately 18x24").
      9.2.4.2  Pole for sampling box (1-1/4" galvanized  steel pipe nominally 10 feet long).
      9.2.4.3  Mounting hardware for sampling box.
      9.2.4.4  Vinyl tape  to cover steel pipe and mounting  hardware.
      9.2.4.5  Trap heater, custom made.
      9.2.4.6  Variable voltage transfer for trap heater.
      9.2.4.7  Vacuum pumps, oil-less, brush-less, capable of flow rates of 30 Lpm and 0.3 Lpm with mass
flow controller.
      9.2.4.8  Rotameters, calibrated, capable of measuring 0.3 and 30 Lpm.
      9.2.4.9  Dry test meter to measure total volume  of air for particulate mercury  sample.
   9.2.5  Freezer for particulate Hg sample storage at -40°C.
Page 5.0-8             Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                       Method IO-5.0
Mercury	Vapor and Particulate Mercury

9.3 Materials

   9.3.1 Supplies for Cleaning Sampling and Analysis Equipment
      9.3.1.1  Acetone, reagent grade.
      9.3.1.2  Alkaline detergent.
      9.3.1.3  Deionized water, high purity, 18.2 M Q cm"1.
      9.3.1.4  Concentrated hydrochloric acid, trace metal grade.
      9.3.1.5  Concentrated nitric acid, trace metal grade.
      9.3.1.6  Water bath capable of maintaining temperature of 80°C.
      9.3.1.7  Polyethylene or polypropylene tubs with lids, various sizes.
      9.3.1.8  Polyethylene carboys, 20 L with spigot.
      9.3.1.9  New, resealable  polyethylene bags.
      9.3.1.10 Particle-free wipes.
   9.3.2 Supplies for Preparation of Gold-Coated Bead Traps
      9.3.2.1  Sputter coater with 24 karat gold source.
      9.3.2.2  Glass-blowing torch.
      9.3.2.3  Tungsten rod for indenting quartz tubes.
      9.3.2.4  Borosilicate glass beads (1 mm dia.), acid-washed.
      9.3.2.5  Quartz tubes (10 cm length,  7 mm O.D.,  5 mm I.D.).
      9.3.2.6  Quartz wool.
      9.3.2.7  Teflon® heat shrink tubing.
      9.3.2.8  Teflon® end plugs, custom made.
      9.3.2.9  Rigid plastic tubes with endcaps for gold  trap storage.
      9.3.2.10 Teflon® tape, Vz".
      9.3.2.11  Polyethylene tube bags and heat sealer.
   9.3.3 Supplies for Preparation of Glass Fiber Filters
      9.3.3.1  Glass fiber filters, 47-mm.
      9.3.3.2  Ceramic crucible.
      9.3.3.3  Teflon® jars for storage of filters.
      9.3.3.4  Teflon® tape, 1 in.
   9.3.4 Supplies for Sample  Collection
      9.3.4.1  Teflon® filter holders, 47-mm, closed inlet for vapor-phase Hg, open inlet for particle-phase
Hg.
      9.3.4.2  Plastic petri dishes, 50 mm,  for particulate mercury sample filters.
      9.3.4.3  Plastic/sealable refrigerator type containers.
      9.3.4.4  Teflon® coated forceps.
      9.3.4.5  Particle-free gloves.
      9.3.4.6  Resealable polyethylene bags.
      9.3.4.7  Teflon® tape, 1"  and Vz".
      9.3.4.8  Teflon® tubing, V4" O.D.
      9.3.4.9  Latex tubing, 1/8" I.D.
      9.3.4.10 Polyethylene tubing, 3/8" O.D.
      9.3.4.11  Quick connectors.
      9.3.4.12 Extension cords.
June 1999              Compendium of Methods for Inorganic Air Pollutants              Page 5.0-9

-------
Method IO-5.0                                                                     Chapter IO-5
Vapor and Particulate Mercury	Mercury

   9.3.5  Supplies for Sample Analysis
      9.3.5.1 Carrier gas, argon or helium, ultra-high purity (99.999%).
      9.3.5.2 Purge gas, nitrogen, pre-purified (99.998%).
      9.3.5.3 Gas tight syringe, 100//L.
      9.3.5.4 Injection port.
      9.3.5.5 Minnert valve and septa for injection port.
      9.3.5.6 Glass impinger and bubblers, 30 mL.
      9.3.5.7 Pipets, including 1000 p.L, 500 p.L, 250 p.L, 100 p.L, and 50 p.L.
      9.3.5.8 Glass or Teflon® tubes, Vz" O.D., 6" long, for soda lime traps
      9.3.5.9 Teflon® reducing unions (Vz" to V4"), for connecting soda lime traps to bubbler exhaust and
gold trap.
      9.3.5.10  Concentrated nitric acid (HN03), highest purity.
      9.3.5.11  Concentrated hydrochloric acid (HC1), highest purity.
      9.3.5.12  Potassium bromide (KBr).
      9.3.5.13  Potassium bromate (KBr03)
      9.3.5.14  Hydroxylamine hydrochloride  (NH2OH-HC1)
      9.3.5.15  Stannous chloride  dihydrate (SnCl2-2H20).
      9.3.5.16  1000 )jg/mL mercury standard, NIST-3133.
      9.3.5.17  Elemental mercury, liquid, high purity.
      9.3.5.18  Constant temperature (± 0.1 °C), circulating water bath.
      9.3.5.19  Certified immersion thermometer.
Page 5.0-10            Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                    Method IO-5.0
Mercury	Vapor and Particulate Mercury
10. Preparation of Supplies, Adsorbents and Reagents

10.1 Acid Cleaning Procedure

[Note:  All Teflon® and polyethylene sample bottles, reagent bottles and analytical supplies which will come
into contact with the samples are cleaned using the following procedures.]

[Note:  The acids mentioned below may be reused many times. A large volume of each type of acid may be
prepared and stored in dedicated polyethylene carboys.  Acid should be checked for contamination or
replaced regularly.]

   10.1.1  Acetone Rinse-sampling and analytical supplies for acid-cleaning by first rinsing with reagent
grade acetone in a fume hood.
   10.1.2  Detergent Wash-supplies are then washed with a dilute Alconox detergent solution, rinsed five
times with cold tap water followed by three rinses with deionized water.
   10.1.3  Heated Hydrochloric Acid Soak-bottles are filled with 3M (250 mL of high purity concentrated
HC1 in 750 mL of D.I. water) trace metal grade HC1 and soaked at 80°C in a water bath in a fume hood for
six hours, then cooled to room temperature, emptied and rinsed three times with ASTM Type I water.
Polyethylene or polypropylene tubs with lids are used to soak small items such as the filter packs, quartz
tubes, Teflon® endplugs, etc.
   10.1.4 Short Nitric Acid Soak-bottles and polyethylene tubs of small items are filled with 0.56M (35 mL
of high purity HN03 diluted to 1,000 mL) trace metal grade HN03, soaked at room temperature for 72 hours,
emptied, and rinsed three times with ASTM Type I water.
   10.1.5  Long Nitric Acid Soak-bottles and covered tubs with supplies are transferred to the Class  100
clean room, rinsed three times with ASTM Type I water, and filled with 0.56 M HN03 and soaked at room
temperature for 7 days. After the soaking period, the bottles & tubs are emptied and rinsed  five times with
ASTM Type I water.
   10.1.6  Drying Step-all sampling and analytical supplies are dried  in the clean room or  in a Class  100
acrylic drying cabinet and double or triple bagged before being stored.
   10.1.7 Bulk Nitric Acid Soak-supplies that cannot be heated in the HC1 step are soaked in covered tubs
with separately prepared 0.56M HN03 at room temperature for 7 days,  rinsed and dried as described above.

10.2 Preparation of Gold-Coated Bead Traps

[Note:  The following sections detail the steps for construction of gold-coated bead traps for  collection of
vapor-phase Hg in field studies.  Depending upon the scope of the mercury sampling to be performed, it may
be cost-effective to purchase traps or trap-making supplies from a commercial source.]

   10.2.1  Preparation of Gold-Coated Beads and Trap
      10.2.1.1  Borosilicate glass beads (1 mm diameter) are coated with a gold plasma generated under
vacuum using  a sputter coater.  The  thickness of the gold coating  produced  should  be  approximately
300 angstrom.  The operating manual for the sputter coater should be consulted for the appropriate settings
to obtain this thickness.
June 1999              Compendium of Methods for Inorganic Air Pollutants            Page 5.0-11

-------
Method IO-5.0                                                                       Chapter IO-5
Vapor and Particulate Mercury	Mercury

      10.2.1.2  Quartz tubes are indented using a glass-blowing torch.  Three radial indentations are made
about 2.5 cm from one end of a 10 centimeter long tube.
      10.2.1.3 All components of the gold bead traps are heated in a muffle furnace to remove any Hg prior
to assembling traps.  The gold-coated beads are baked at 500°C for one hour, the quartz wool at 600 °C for
one hour, and the quartz tubes at 700°C for one hour.
      10.2.1.4 Teflon® heat-shrink tubing (0.2 5 in. diameter after shrinking), is cut into 1.2 5 in. pieces and
acid cleaned with a bulk HN03 soak (see Section 10.1.7).
   10.2.2  Construction of Trap
      10.2.2.1  A small amount of quartz wool is rolled into  a ball and placed in the quartz tube so that it
rests on the indentations forming a short (~ 0.25 cm) plug.  About 0.7 g of the gold-coated beads are added
to the quartz tube on top of the quartz wool.  A second ball of quartz wool is placed in the quartz tube so that
it rests on top of the beads. Traps should be packed as tightly as possible without breaking the quartz wool
fibers. Three more radial indentations in the quartz tube are made just beyond the second quartz wool plug.

[Note: Best results are obtained by maintaining the integrity of the quartz wool fibers when rolling and
compacting the wool into the quartz tube. Also, loosely packed traps may result in air flowing around the
beads and not contacting the gold surfaces.]

      10.2.2.2 Teflon® heat-shrink tubing is attached to both ends of the trap so that about half of the tubing
extends from the trap.  The Teflon® endplugs are inserted into the Teflon® heat-shrink tubing.
      10.2.2.3  Plastic tubes are used as storage containers for each individual trap with endplugs.
      10.2.2.4  Each trap is labeled with a unique number to identify and monitor the performance of the
trap.
   10.2.3  Trap Conditioning and Testing
      10.2.3.1  New traps are  'conditioned' prior to use by drawing air through the trap at 0.3 Lpm for 30
minutes and then 'blanked' by heating the trap to 500°C for five minutes while an inert gas (i.e., He) flows
through the trap at 0.3 Lpm. The conditioning procedure is performed twice prior to testing the trap.
      10.2.3.2  New traps are evaluated for their reproducibility in collecting vapor-phase Hg and for their
blank levels prior to use.
      10.2.3.3  Traps to be used in vapor-phase Hg sample collection are prepared by blanking the  trap,
Teflon®-taping the endplugs to the  trap, placing the trap in  the plastic storage tube, heat sealing the tube in
polyethylene, then placing two gold traps for a sample in a  resealable polyethylene bag.

10.3 Preparation of Glass Fiber Filters

[Note: Glass fiber filters used for collection of particle-phase Hg samples must be heated prior to use to
release any Hg from  the filter matrix.]

   10.3.1  The glass  fiber filters are placed in a clean crucible with a lid and heated in a muffle furnace at
500 °C for one hour.
   10.3.2  After cooling for one hour in the muffle furnace  but still hot, the filters are transferred from the
crucible to an acid cleaned Teflon® jar (50 mm diameter) using acid cleaned Teflon® forceps. The lid to the
Teflon® jar is sealed with Teflon® tape, and the jar is triple-bagged and stored in the clean room until needed.
   10.3.3 Twenty percent of each batch of glass fiber filters used for particle-phase Hg samples and blanks
are  kept for preparation of standard addition filters used for CVAFS instrument calibration  as described in
Section 12.6.2.
Page 5.0-12             Compendium of Methods for Inorganic Air Pollutants              June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury	Vapor and Particulate Mercury
11. Collection of Ambient Air Samples

[Note: A basic concern regarding sample collection is the potential for contamination.  Absolute adherence
to clean sampling protocols is essential.  This includes, but is not limited to:  (1) all supplies with which
samples come into contact must be acid cleaned, (2) all sample containers must be handled with particle-free
gloved hands at all times, (3) all sample containers must be bagged before and after sample collection, and
(4) the operator must stand downwind of the sample during all phases of sample installation and removal.

In addition, all sample collection procedures (i.e. loading filters into filter packs) should be conducted
outdoors (in the case that a Class 100 laboratory is unavailable on-site). If atmospheric mercury sampling
is to be performed in locations that experience extreme weather conditions, additional measures can be taken
to provide a 'clean' environment indoors for sample handling. Small laminar-flow hoods in site trailers have
been used with good results for loading and unloading sample filters from filter packs.

This section describes the methods for collection of vapor and particle phase Hg samples.  The configuration
of the sampling equipment is described first followed by the specific procedures for collection of vapor and
particle phase Hg, respectively.  In the following, the term 'gold trap' is used to refer to traps made from
gold-coated beads, gold-coated sand and/or a solid gold matrix which have been shown to produce equivalent
results.]

11.1 The Sampling Equipment

   11.1.1 The pumps used for collection of mercury samples in  ambient air should  be specially designed for
trace-level pollutant sampling.  High efficiency oilless, brushless pumps should be used and protected from
weather (i.e., the pump housing should be well sealed from  rain, insulated and heated during the winter, and
fan cooled during the summer).
   11.1.2  A fiberglass enclosure is used as the 'sample box' to house the filter packs and gold traps during
sample collection (see Figure  1).
   11.1.3 The box  is mounted on a pole or tower at least 3 meters  above ground level.  Galvanized steel
provides  sufficient strength to support the sample box in high winds and may be wrapped with vinyl tape to
prevent contamination from the metal surface.
   11.1.4 Polyethylene tubing with quick connect couplings are used to connect the vacuum lines from the
pump to the sample  box.
   11.1.5 For particle-phase mercury, a dry test meter (DTM)  is connected to the vacuum line between the
pump and the sample box for direct measurement of the volume of air sampled (in L).
   11.1.6 Inside the sample box flexible latex tubing with  quick connect couplings are used to connect the
vacuum lines to the filter packs for particle-phase Hg.   For vapor-phase Hg a piece of Teflon® tubing is
attached to the end of the flexible latex tubing for connecting to the gold traps during sampling.
   11.1.7 Moisture on gold surface interferes with the amalgamation of Hg. Therefore,  condensation during
collection of vapor-phase Hg must be prevented. To prevent condensation in the gold traps during sampling,
a heat tube is used to keep the traps above ambient temperature.  The heat tube is constructed from heat tape
wrapped  around a metal tube, insulated and covered with heat-shrink electrical tubing and electrical tape.
The heat tube assembly is plugged into a variable voltage transformer to provide constant low heat to warm
the trap.
June 1999             Compendium of Methods for Inorganic Air Pollutants            Page 5.0-13

-------
Method IO-5.0                                                                      Chapter IO-5
Vapor and Particulate Mercury	Mercury

   11.1.8 The 47-mm Teflon® filter packs used for collection of vapor- and particle phase Hg are an
assemblage of three main components—a threaded inlet, a filter support base with V4 inch tube ferrule nut,
and a clampdown nut to connect the inlet to the filter holder.  For vapor-phase Hg a closed inlet filter pack
with approximately 4" of ¥4 inch Teflon® tubing is used to extend the air inlet several inches from the bottom
of the sample  box.  For particle phase Hg an open inlet filter pack is used that protrudes several inches
through the sample box during sampling and a plastic quick connect coupling is used for connection of the
outlet to the vacuum line.
   11.1.9 Rotameters calibrated for 0.3 and 30 Lpm are used initially to set the flow rate for vapor and
particle-phase  Hg respectively. To prevent potential contamination of the sample by the rotameter, separate
'flow check' filter packs are attached to the rotameters for this procedure. The flow check filter packs have
closed inlets with quick connect couplings on the inlet tubing for connection to the rotameter.
   11.1.10 A  field test data sheet (FTDS) is used to record all field activities.

11.2 Collection of Vapor Phase Hg Samples

[Note: Collection of vapor-phase mercury in ambient air takes advantage of the amalgamating property of
mercury to a gold surface. The amalgamation process requires a low flow rate to allow adsorption of the
mercury in the air to the gold surface of the trap.  A nominal flow rate of 0.3 Lpm is used to collect vapor
phase Hg onto gold traps.  A mass flow controller is employed to ensure a constant flow rate throughout the
sampling period and the flow rate is checked both at the beginning and at the end of the sampling period with
a calibrated rotameter.  The volume of air sampled is determined from the average  of the flow rate
measurements and the sample duration.  Checks of the flow rate also test for leaks and obstructions in the
vacuum lines.

Vapor phase Hg samples are collected using two gold traps in series as a quality control measure.  The outlet
from the first trap ('A' or 'sample' trap) is connected to the inlet of the second trap ('B' or 'back' trap) using
Teflon® tubing.  Any breakthrough from the first trap is collected on the second  trap. The two traps are
attached to a Teflon® filter pack containing a glass fiber filter to prevent intrusion of particulate matter into
the gold traps.

   11.2.1  Flow check procedure
      11.2.1.1 The flow rate through the gold traps is set to 0.3 Lpm before each sample event using a
rotameter calibrated for 0.3 Lpm.  To prevent potential contamination of the sample by the rotameter,  a
separate flow check filter pack and gold trap are attached to the rotameter for this procedure. The flow check
glass fiber filter should be replaced regularly but may be reused several times.
      11.2.1.2 The sampling pump is allowed to warm up for at least  15 minutes prior to any flow
measurement and/or adjustment.
      11.2.1.3 The  flow check filter pack and gold trap are placed in the sampling box and the vacuum line
connected to the gold trap. The rotameter is connected to the  inlet of the filter pack.  After the system has
been allowed to stabilize,  the flow rate is adjusted to 0.3 Lpm  and is recorded on the FTDS. The rotameter
is disconnected from the filter pack and the gold trap disconnected from the vacuum line.  The flow check
filter pack and gold  traps  are removed from the sampling box.
Page 5.0-14             Compendium of Methods for Inorganic Air Pollutants              June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury _ Vapor and Particulate Mercury

   11.2.2 Sample installation procedure

[Note: During all phases of sample set-up and removal, the operator stands downwind of the sample in order
to prevent contamination by shedding particles from clothing, etc. In addition, particle-free gloves are worn
when handling gold bead traps and prefilters.]

      11.2.2.1  An acid cleaned 'sample' filter pack with closed inlet is loaded with a pre-fired glass fiber
filter for each new vapor phase Hg sample to be collected and placed in the sample box.
      11.2.2.2  The gold traps are taken from their plastic storage  tubes and the endplugs removed.
      11.2.2.3  The 'A' trap is connected to the Teflon® tubing from the outlet of the filter pack. A short
piece of acid-cleaned Teflon® tubing is placed in the outlet of the 'A' trap.  The 'B' trap is then attached to
the Teflon® tubing.
      11.2.2.4  The heat tube assembly is positioned to cover the 'A' trap.
      11.2.2.5  To begin the sample, the Teflon® tubing at the end of the vacuum line is attached to the outlet
of the 'B' trap.
      11.2.2.6  Record the following parameters on the Hg FTDS:  date, sampling location, time, ambient
temperature, barometric pressure, relative humidity, mass flow control reading, rotameter reading, and gold-
trap  number.  A typical Hg FTDS is documented in Figure 2.
      11.2.2.7  Allow the sampler to operate for the desired time. At the end of the sampling period, record
the parameters listed in Section 11.2.2.6.  If the flows at the beginning and end of the sampling period differ
by more  than 10 percent, mark the gold-trap cartridge as suspect.
      11.2.2.8  Calculate and record the average sample rate for the gold-trap cartridge according to  the
following equation:
                                           ,
                                                  N

where:

              Qa =  Average flow rate in L/minute.
           Q!, Q2 =  Flow rates determined at beginning and end points during sampling.
               N=  Number of points averaged.
      11.2.2.9 Calculate and record the total volumetric flow for the gold-trap cartridge using the following
equation:
                                                (T)(Qa)
                                                 1,000
where:
  Vm =  Total volume sampled (m3) at measured temperature and pressure.
    T =  Sampling time = T2 - Tj, minutes
   T2 =  Stop time.
June 1999             Compendium of Methods for Inorganic Air Pollutants            Page 5.0-15

-------
Method IO-5.0                                                                     Chapter IO-5
Vapor and Particulate Mercury	Mercury

   Tj =  Start time.

      11.2.2.10  The total volume  (Vs) at standard conditions, 25 °C and 760 mmHg, is calculated from the
following equation:
                                              PA       298
                                           X —— X
                                              760    273  +  tA
where:
   Vs = Total sample volume (m3) at standard conditions, 25 °C and 760 mmHg.
   PA = Average barometric pressure, mmHg.
   tA = Average ambient temperature, °C.

   11.2.3  Sample recovery procedure

[Note:  Particle-free gloves are worn during this procedure.]

      11.2.3.1  The gold-traps are removed from the sampling box and the endplugs are replaced.  The
endplugs are sealed to the traps with Teflon® tape and the traps are returned to their respective plastic tubes
and capped.
      11.2.3.2  Labels with unique sample identification numbers are  affixed to each tube.  The tubes are
sealed in polyethylene bags and returned to the laboratory for analysis.
      11.2.3.3 The flow rate is checked after removal of the sample traps and filter pack using the procedure
previously described in  Section 11.2.1.  The flow  rate and any deviations from the standard  operating
procedures during removal of the sample must be recorded on the Hg  FTDS.
      11.2.3.4  The  pump is  turned off and the glass fiber prefliter is  discarded.  The Teflon® tubing
connector should be returned to the laboratory for cleaning.

11.3 Collection of Particulate Hg Samples

[Note:  Since particulate Hg occurs at ultra-trace levels in the atmosphere and Hghas a high vapor pressure,
the selection of sampling flow rate and duration must be carefully considered.  It is typically necessary to
sample at a flow rate of 30 Lpm for 12-24 hours to collect enough particulate Hg for analysis.  The volume
of air sampled is measured directly using a calibrated dry test meter (DIM).  In addition, the flow rate is
checked both at the beginning and at the end of the sampling period with a calibrated rotameter to check for
leaks or obstructions in the vacuum lines.]

   11.3.1  Flow check procedure
      11.3.1.1 The flow rate is checked before each sample event using a rotameter calibrated in the range
of 30 Lpm. To prevent potential contamination of the sample by the rotameter, a separate flow check filter
pack is attached to the rotameter for this  procedure.  The flow check  glass fiber filter should be replaced
regularly but may be reused several times.
      11.3.1.2  The  sampling pump is allowed to  warm up for at least 15 minutes prior to  any flow
measurement and/or adjustment.
      11.3.1.3  The flow check filter pack is placed in the sampling  box.  The outlet is connected  to the
vacuum line and the inlet to the rotameter.  After the system has been  allowed to stabilize, the flow rate is
Page 5.0-16            Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                    Method IO-5.0
Mercury _ Vapor and Particulate Mercury

adjusted to 30 Lpm and is recorded on the FTDS. The flow check filter pack is then disconnected from the
rotameter and vacuum line, and removed from the sampling box.
   11.3.2 Sample installation procedure

[Note: During all phases of sample set-up and removal, the operator stands downwind of the sample in order
to prevent contamination by shedding particles from clothing, etc. In addition, particle-free gloves are worn
when handling gold bead traps and pre filters.]

      11.3.2.1 An acid cleaned 'sample' filter pack with open cylinder inlet is loaded with a pre-baked glass
fiber filter with the 'fibrous side' up, touching only  the edge of the filter with a pair of acid cleaned Teflon®
coated forceps.  The filter pack is placed in the sample box.
      11.3.2.2 To begin  the sample, the vacuum line is connected to the outlet of the sample filter pack.
      11.3.2.3 Record the following parameters on the Hg Field Test Data Sheet (FTDS):  date, sampling
location, time,  ambient temperature,  barometric  pressure, relative humidity,  dry gas  meter reading,
rotameter reading (if applicable), gold-trap number and dry gas meter serial number. A typical Hg FTDS
is documented in Figure 2.
      11.3.2.4 Allow the sampler to operate for the desired time. At the end of the sampling period, record
the parameters listed in Section 11.3.2.3.  If the flows at the beginning and end of the sampling period differ
by more  than 10 percent,  mark the filter cartridge as suspect.
      11.3.2.5  Calculate and record the average sample rate for  the filter cartridge according to the
following equation:
                                         Q  ,L
                                         y-       N

where:

              Qa =   Average flow rate in L/minute.
           Qj, Q2 =   Flow rates determined at beginning and end points during sampling.
               N=   Number of points averaged.
      11.3.2.6 Calculate and record the total volumetric flow for the gold-trap cartridge using the following
equation:
                                                (T)(Qa)
                                                 1,000
where:
  Vm =  Total volume sampled (m3) at measured temperature and pressure.
    T =  Sampling time  = T2 - Tj, minutes
   T2 =  Stop time.
   T, =  Start time.
June 1999              Compendium of Methods for Inorganic Air Pollutants            Page 5.0-17

-------
Method IO-5.0                                                                    Chapter IO-5
Vapor and Particulate Mercury	Mercury

      11.3.2.7  The total volume (Vs) at standard conditions, 25 °C and 760 mmHg, is calculated from the
following equation:
where:
                                              760    273 + tA
   Vs = Total sample volume (m3) at standard conditions, 25 °C and 760 mmHg.
   PA = Average barometric pressure (mmHg) during sampling.
   tA = Average ambient temperature (°C) during sampling.
   11.3.3  Sample recovery procedure

[Note:  Particle-free gloves are worn during this procedure.]

      11.3.3.1 The  sample filter is removed from the filter pack and placed in an acid cleaned petri dish
using acid-cleaned Teflon®-coated forceps, touching only the edge of the filter.
      11.3.3.2 The petri dish is sealed with Teflon® tape and a label with a unique sample identification
number is affixed to the cover of the petri dish. The petri dish is double bagged and returned to the laboratory
for analysis.
      11.3.3.3 The flow rate is checked after removal of the sample Hg traps using the procedure previously
described in Section  11.3.1. The flow rate and any deviations from the standard operating procedures during
removal of the sample must be recorded on the Hg FTDS.

11.4 Sample Storage

   11.4.1  Vapor-phase Hg samples can be stored up to one week at 4°C until analyzed.
   11.4.2  Particle-phase Hg filters can be stored indefinitely in a freezer at -40°C until analyzed.
Page 5.0-18            Compendium of Methods for Inorganic Air Pollutants              June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury	Vapor and Particulate Mercury
12. Analysis of Ambient Air Samples

12.1 Introduction

   12.1.1 This section describes the methods for analysis of vapor and particle phase Hg samples.  Dual-
amalgamation cold vapor atomic fluorescence spectrometry (CVAFS) is used to determine the amount of Hg
collected for both types of samples.
   12.1.2 The analytical system is described first followed by the specific procedures for analysis of vapor
and particle phase Hg, respectively.

[Note:  In the following, the term 'gold trap' is used to refer to traps made from gold-coated beads, gold-
coated sand and/or a solid gold matrix which have been shown to produce equivalent results.]

12.2 The Analytical System

   12.2.1 The analytical system consists of a pair of gold traps in series, a thermal desorption system, a
CVAFS mercury detector, an integrator to convert peak voltage to peak area,  and an inert carrier gas (He
or Ar).
   12.2.2  The dual-amalgamation technique  requires two gold  traps in series: a 'sample'  trap and  an
'analytical' trap.  This technique has two main advantages:   it virtually eliminates interferences due to
organics  and C12 from the sample trap and provides greater analytical precision because the same analytical
trap is being used to introduce the Hg into the detector for all  the samples.
   12.2.3 The thermal  desorption system includes a pair of nickel-chromium coils each with a variable
transformer to  regulate the output, a pair of small axial fans to  cool  the coils and traps  quickly,  and a
programmable circuit controller to coordinate the trap heating and cooling cycles.
   12.2.4 Power supplied to the CVAFS should be stabilized by a power conditioner to reduce line voltage
fluctuations. The instrument is left on continuously to stabilize the UV lamp and maintain consistency from
one day to the next.  Operating manuals for the CVAFS instruments describe in detail the operation and
maintenance for the specific analyzer.
   12.2.5 The carrier gas should be mass-flow controlled to produce a more consistent peak response.  Ultra-
high purity gas should be used and a gold trap placed upstream of the sample and analytical traps to remove
any remaining traces of Hg in the gas.
   12.2.6 The vapor phase Hg analytical system requires an injection port and gas-tight syringe (100 //L)
to introduce Hg standards into the sample train.
      12.2.6.1  A  small amount (2-3 mL) of liquid metallic Hg in a closed 150  mL flask is sufficient to
generate  Hg saturated air above the liquid surface for vapor-phase  Hg analysis.  The flask is immersed in a
constant temperature (± 0.1 °C) circulating water bath.  A certified immersion thermometer should be used
to monitor the temperature of the air above the mercury in the flask.  The temperature of the air in the flask
must be maintained below room temperature otherwise, the Hg will condense on the walls of the injection
apparatus.
      12.2.6.2  The gas-tight  100  /jL  syringe should rest on top  of the flask  with  the needle portion
protruding through a Minnert valve into the flask air.  Hg saturated air  from the flask should be drawn  up
into the syringe and allowed to equilibrate.
June 1999              Compendium of Methods for Inorganic Air Pollutants            Page 5.0-19

-------
Method IO-5.0                                                                     Chapter IO-5
Vapor and Particulate Mercury	Mercury

      12.2.6.3 An injection port with Minnert valve and Teflon-coated septum are used for injecting Hg
saturated air onto gold traps in the analytical system. The port is placed inline for the calibration process and
then removed.
      12.2.6.4 Vapor phase analysis is done outside of the clean room.
   12.2.7 The particle-phase Hg analytical system requires an aqueous purging system which consists of:
(i) an acid cleaned glass impinger and bubbler system (30 mL capacity) with Teflon® stopcock,  (ii) a  soda
lime trap to capture acid gases from the bubbling solution,  (iii)  gold traps for collection of volatilized Hg
from the  extracts, and (iv) N2 carrier gas for the system.

12.3 Preparation of Reagents and Standards

   12.3.1 Deionized water-Deionized water, with a resistivity of 18.2  MQ cm, is prepared using a water
purification system (e.g., Milli-Q)  from a prepurified (reverse osmosis) water source.  This purified water
conforms to the standards for ASTM Type I water.
   12.3.2 Hydrochloric acid-containing low concentrations of Hg are used to prepare BrCl and SnCl2
solutions. EM Science SupraPur®  HC1 is recommended.
   12.3.3 Bromine Monochloride-(BrCl) is made from high purity potassium bromide (EM Science) and
potassium bromate (Baker).  The bromine salts (10.8 g of KBr and 15.2 g of KBr03) are weighed using acid
cleaned weigh boats and spatulas.  In the clean room laminar flow exhaust hood, an acid cleaned magnetic
stir bar is placed  in a 1L glass bottle of concentrated, high purity HC1.  The KBr is added while stirring the
acid with a magnetic  stir plate,  and allowed to dissolve completely (approx. 1 hour).   The KBr03 is  then
added slowly  while stirring. After all of the salts have been added, the solution is allowed to mix until all
visible particles have dissolved (1-2 hours). The solution should be deep yellow in color. If there is no color
(or very faint) then the solution should be remade because the BrCl has undergone reduction and will not have
sufficient strength to  oxidize all mercury species to Hg+2.  The BrCl is stored at room temperature in the
clean room exhaust fume hood.

[Caution: BrCl should always be prepared in an exhausting fume hood because hazardous chlorine (CIJ gas
is produced.]

   12.3.4 Hydroxylamine hydrochloride-(NH2OH-HCl) solution  is prepared by dissolving 75.0  g of
NH2OH-HC1 (EM Science) in ASTM Type I water in an acid cleaned 250 mL volumetric flask. The solution
is stored  in an opaque Teflon® bottle and refrigerated when not being used.
   12.3.5 Stannous chloride-(SnC12) solution is prepared by placing 200.0 g of SnCl2-2H20 (Fluka) in an
acid cleaned 1000 mL volumetric flask.  In the clean room laminar flow exhaust fume hood,  100 mL of
concentrated,  high purity HC1 is slowly added.  After most of the  SnCl2 has dissolved  (it does not dissolve
completely), the solution is brought up to volume with ASTM Type I water.  The SnCl2 solution is stored in
an acid cleaned, opaque Teflon® bottle and refrigerated when not being used.
   12.3.6 Working mercury standard-solution used  for CVAFS instrument calibration is prepared by
sequential dilution of a commercial primary standard  of 1000 p.g Hg/mL.  A secondary standard with  a
concentration of 1000 ng Hg/mL is made by adding 100 p.L of the primary standard and 5 mL of BrCl (see
Section 12.3.3) to a  100 mL volumetric  flask, and diluting to the 100 mL total volume with ASTM Type  I
water.  The secondary standard solution is remade  at least once per year. The working standard has  a
concentration of 2 ng Hg/mL in 1 percent BrCl and is prepared by adding 200 p.L of the secondary standard
and 1000 p.L of BrCl to a 100 mL volumetric flask and  diluting to  100 mL total volume with ASTM Type  I
water.  All standard solutions are stored in the dark at 5°C.
Page 5.0-20            Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                      Method IO-5.0
Mercury	Vapor and Particulate Mercury

   12.3.7 Nitric acid-(HN03) extraction solution (10% HN03, 1.6 M) is used for microwave digestion of
particulate Hg samples. In the clean room laminar-flow exhausting hood 100 mL of high purity concentrated
HN03 (EM Science SupraPur®) is slowly added to 750 mL of ASTM Type I water in a 1000 mL volumetric
flask. The solution is mixed, allowed to cool, and brought up to 1000 mL with ASTM Type 1 water.  The
extraction solution is stored in an acid cleaned repipetting bottle in the clean room exhaust hood.

12.4  Summary of Dual-amalgamation CVAFS Analytical Procedure

   12.4.1  The analytical procedure for detection of mercury using CVAFS includes three main steps.
   12.4.2  First, the sample trap is placed in the  analytical  system and heated  to release the collected
mercury.  The released mercury is entrained by the gas stream, carried into the analytical trap where it is
amalgamated to the gold surface. Second,  the analytical trap is heated to release the mercury which  then
flows into the CVAFS detector cell.  Third, the mercury in  the detector cell absorbs UV light and the
resulting fluorescence is converted to a  voltage  proportional to  the  amount  of Hg  detected  by a
photomultiplier tube.
   12.4.3 An integrator then converts the voltage response to  peak area.

12.5  Analysis of Vapor Phase Hg Samples

   12.5.1 Injection  system calibration
      12.5.1.1  The  injection system  is conditioned before each day of analysis to ensure  precise and
reproducible results.  The process includes the following steps:  (i) condition the syringe by flushing it three
times with Hg saturated air,  filling it to capacity, and allowing it to equilibrate for 15 minutes, (ii) place a
new septum in the injection port and condition them both with multiple injections of Hg saturated  air before
initiating calibration.
      12.5.1.2  The flask should  be periodically flushed (approx. once per month) with pre-filtered N2 gas
in order to displace any oxygen which may oxidize the surface  of the  liquid mercury and prevent
volatilization.  Also,  the Minnert valve  for the syringe on the flask should be above the level of the water in
the water bath.  If water gets into the flask,  it should be purged as described above.
   12.5.2 CVAFS instrument calibration
      12.5.2.1  A calibration curve, generated by injection of different volumes of Hg saturated  air onto a
gold trap, must be performed before each analysis.   The amounts of Hg injected for the calibration curve
should be tailored to the expected value of the samples to be analyzed.  Table 1 shows an example calibration
curve for 24 hour vapor-phase Hg samples consisting of five different amounts of Hg injected:  0, 0.2, 0.4,
0.8 and  l.GngofHg.
      12.5.2.2  To generate standard injections for the calibration curve, the  conditioned injection port is
placed in the analytical train  in front of the gold trap to be used  for generating the standard curve (called the
"standard trap").  A  specified volume  of Hg saturated air is withdrawn from the flask using the gas tight
syringe and injected onto  the blanked standard trap through the injection  port valve. After the injection the
syringe is returned to the flask and filled to capacity  until the next injection.  The temperature of the air in
the flask is recorded for each injection to calculate the actual amount of Hg injected.
      12.5.2.3  A calibration curve is  performed beginning with a zero  point (0 ,uL) and continuing in
ascending order to the highest amount.  The zero point of the calibration curve is generated in the  same
manner described above except that no  Hg saturated air is injected into the port. The amount of Hg emitted
from  the needle tip and the injection apparatus and adsorbed onto the  gold  trap  is called the zero  point
(typically between 1-5 pg Hg).
June 1999              Compendium of Methods for Inorganic Air Pollutants            Page 5.0-21

-------
Method IO-5.0                                                                      Chapter IO-5
Vapor and Particulate Mercury	Mercury

      12.5.2.4 The response for each standard injection is obtained by dual-amalgamation CVAFS described
in Section 12.2.
      12.5.2.5  At any given temperature the vapor density of Hg can be calculated using the Ideal Gas Law
and the saturation vapor pressure of Hg.  A table of vapor densities versus temperature is used to determine
the amount of Hg injected for the volume of each standard injection.  Table 1 illustrates the amount of Hg
delivered during a typical injection procedure used to generate a standard curve  at a flask temperature of
16.6°C.
      12.5.2.6  The slope of the calibration  curve is calculated using linear regression.  The 0 ,uL injection
response is subtracted from  each of the points on the curve.  The CVAFS analytical system is linear and the
coefficient of determination (r2) should be 0.999 or better and each of the points on the curve should be
predicted by the slope within 5 percent of their actual value.  If these criteria are not met, the specific points
that are errant should be repeated and the linear regression recalculated.
      12.5.2.7  The injection port and standard trap are removed from the analytical train after calibration.
   12.5.3  Sample Analysis
      12.5.3.1 After a satisfactory calibration curve for the CVAFS is obtained, sample traps are analyzed
by dual amalgamation CVAFS as described  in Section 12.2.
      12.5.3.2  Control standards should be analyzed at regular intervals during the analysis of samples to
detect any drift in the response or change in sensitivity  of the  CVAFS instrument.  Control standards are
generated in the same manner as the calibration standards described above.  The volume of Hg saturated air
injected for a control standard should be representative of the Hg  concentration of the samples being
analyzed.
      12.5.3.3  The predicted value of the control standards should be within 5% of the calibration curve
before continuing to analyze samples.  If subsequent control standards also deviate by more than 5% from
the calibration curve, it  is likely that the analyzer sensitivity has changed and  a  second calibration curve
should be generated.  Sample  analysis should only continue after recalibration has met the criteria  outlined
in 12.5.2.6.

12.6 Analysis of Hg in Particulate Samples

   12.6.1  Microwave Digestion System
      12.6.1.1  The  use of  microwave digestion to extract samples for analysis  allows rapid heating and
elevated pressures for shorter preparation time in a safer and more consistent manner compared to traditional
procedures.
      12.6.1.2 The microwave digestion system must use Teflon® vessels and be capable of monitoring and
controlling pressure and temperature inside  the vessels.
      12.6.1.3  The microwave should be programmable for the target  pressure and temperature, and for
maintaining them for a specified time.  The operating manual for the microwave digestor should be consulted
for programming, safety and maintenance of this equipment.
   12.6.2  Preparation of Standard Addition  and Sample Filters for Extraction

[Note: All preparation of filters for particulate Hg analysis must be performed in a Class 100 clean room.
Filters are removed from storage in a -40 °C freezer and allowed to equilibrate  to room temperature in the
clean room. Acid-cleaned Teflon® microwave vessels and Teflon®-coated forceps are needed for the filter
preparation.  Calibration standards using filter media are prepared at the same time as the sample filters to
be analyzed. The microwave vessels should be labeled with the appropriate identifier for the contents of each
vessel]
Page 5.0-22             Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                      Method IO-5.0
Mercury	Vapor and Particulate Mercury

      12.6.2.1  Glass fiber filters heated and stored with the sample filters as described in Section 10.3 are
used as standard addition filters for calibration of the CVAFS instrument.  In the clean room, the standard
filters are placed into acid cleaned microwave vessels using acid cleaned Teflon®-coated forceps. The filters
are spiked with appropriate volumes of the 2 ng/mL Hg working standard to generate a calibration curve.
Table 2 shows the volume of Hg standard added to the filters in the microwave vessels to produce a typical
calibration curve for 24 hour particulate Hg samples. The Hg working standard solution is pipetted directly
onto the filter and absorbed by the filter.
      12.6.2.2  All sample filters to be analyzed should be prepared at the same time and handled identically
according to the following procedures. Each petri dish containing a sample filter is removed from the sealed
polyethylene bag and the Teflon® tape around  the dish removed and discarded.  The filter is removed from
the petri dish using acid cleaned Teflon®-coated forceps touching only the edge of the filter to avoid the center
area with the collected particulate matter.  The filter is carefully  folded into quarters using two pairs of
forceps and placed into the appropriately labeled vessel. The forceps are cleaned after touching each sample
filter by rinsing the tips in acid cleaned glass beakers of 10% HN03 extraction solution and ASTM Type  I
water.
   12.6.3 Filter Extraction using Microwave Digestion

[Note:  The  prepared standard and sample filters are handled identically using the following microwave
extraction procedure.]

      12.6.3.1 In the  clean  room exhaust hood, the extraction solution (10% HN03, 1.6 M) should be
prepared as  described in Section 12.3.  A calibrated repippetor is used to dispense 20  mL of extraction
solution into each vessel.
      12.6.3.2  Each vessel should be weighed both prior to and following the microwave digestion to ensure
no loss of sample extract during the procedure.
      12.6.3.3  The prepared vessels can then be removed from the clean room and placed in the microwave
oven for  digestion.  The pressure and temperature monitoring sensors should be appropriately attached and
the control program initiated.  Optimal results for digestion of glass fiber filters for mercury determination
have been obtained  by heating the samples at 160°C and 70 psi for 20 minutes.
      12.6.3.4  After the microwave digestion procedure is complete, the vessels should be allowed to cool
until the pressure inside the control vessel is about 1-2 psi. then removed from the microwave and returned
to the clean  room.
   12.6.4 Oxidation of Digested Filters
      12.6.4.1  In the clean room exhaust hood, the vessels are opened carefully and 0.5 mL of BrCl (see
Section 12.3.3) is added to the extract in each vessel to oxidize all of the mercury in solution to Hg2+.
      12.6.4.2  The capped vessels are gently swirled to mix the vessel contents and allowed to react for  a
least 1 hour.
   12.6.5 Aqueous Purging  System

[Note:  In the clean room exhaust hood an aqueous purging system is used to volatilize Hg° from the filter
extraction solution onto gold traps for quantification by CVAFS.  The flow rate of the carrier gas through the
purging system should be maintained at 0.5 Lpm.]

      12.6.5.1  A soda lime trap is made from an acid cleaned borosilicate glass tube (15 cm long, 15 mm
outer diameter) or from Teflon® tubing of similar dimensions. Acid cleaned Teflon® tubing and compression
fittings are used to attach the trap to the impinger exhaust.
June 1999              Compendium of Methods for Inorganic Air Pollutants            Page 5.0-23

-------
Method IO-5.0                                                                      Chapter IO-5
Vapor and Particulate Mercury	Mercury

      12.6.5.2 Before beginning analysis the tube is packed with high purity grade soda lime using a small
amount of glass wool (1/2") as endplugs. The soda lime trap is then conditioned by purging 20 mL of a 5%
HC1 solution (0.3 M) for 30 minutes.
      12.6.5.3 Purging solution is prepared by adding 1 mLSnC!2 (Section 12.3.5) to 20mLof ASTMType
I water in a bubbler and purging for 15 minutes to remove any Hg  in the solution or purging system.
   12.6.6 Volatilization and Recapture Procedure

[Note: The following volatilization and recapture procedure is used to liberate Hg° from the oxidized filter
extracts and collect it on gold traps for quantification by CVAFS.]

      12.6.6.1 A 5 mL aliquot of the oxidized filter extract is pipetted from the Teflon® vessel into 20 mL
of the bubbling solution prepared above.
      12.6.6.2 To reduce  the halogens in the extract, 0.1 mL NH2OH-HC1  (see Section 12.3.4) is pipetted
into the bubbler.  The bubbler is swirled briefly to mix the solution and allowed to react for 5 minutes.
      12.6.6.3 A blanked gold trap is attached to the end of the soda lime trap connected to the impinger for
collection of mercury from the bubbled solution.
      12.6.6.4 To reduce the oxidized mercury in solution to volatile Hg°, 0.5 mL SnCl2 is pipetted into the
bubbler which is immediately attached to the impinger and purged  onto the gold trap for 7 minutes.
   12.6.7 CVAFS Instrument Calibration using Standard Addition Filters
      12.6.7.1 A calibration curve, generated using the standard addition filter extracts, must be performed
before each analysis. A calibration curve is performed beginning with the reagent blank (no Hg standard
added to  filter) and continuing in ascending order to the highest concentration standard.
      12.6.7.2 The volatilization and recapture procedure described in Section  12.6.6 is used to collect the
Hg in solution for each standard addition filter onto  gold traps.  The amount  of Hg volatilized from the
extracts is determined  by analysis  of the gold trap using  dual-amalgamation CVAFS  as  described in
Section 12.4.
      12.6.1.3 The concentration of Hg in the extraction solution  for each  of the standard addition filters
is determined from the amount of 2 ng/mL mercury standard added to the filter,  the total volume of reagents
added, and the volume of solution analyzed. The actual amount of mercury in 5 mL of standard filter extracts
are as shown in Table 2.
      12.6.1.4 The slope of the calibration curve is calculated using linear regression.  The reagent blank
response is subtracted from each of the points on the curve.  The CVAFS analytical system is linear and the
coefficient of determination (r2) should be 0.99 or  better and each of  the points on  the curve should be
predicted by the slope within 10% of their actual value. If these criteria  are not  met, the specific points that
are errant should be repeated and the linear regression recalculated.
   12.6.8 Sample Analysis
      12.6.8.1 After a satisfactory calibration curve for the CVAFS  is obtained using standard addition filter
extracts,  Hg in each sample filter extract is volatilized and recaptured using  the procedure  described in
Section 12.6.2.
      12.6.8.2 The gold traps are analyzed by dual amalgamation  CVAFS as described in Section 12.4 to
quantify the amount of Hg in the sample filter extract.
      12.6.8.3  Control standards should be analyzed at regular intervals during the analysis of samples to
detect any drift in the response or change in sensitivity of the  CVAFS  instrument.  Control standards are
generated in the same manner as the calibration standards described above. The standard addition filter used
for a control standard should be representative of the Hg concentration of the samples being analyzed.
Page 5.0-24             Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                    Method IO-5.0
Mercury	Vapor and Particulate Mercury

      12.6.8.4  The control standards response should  be within 10% of the calibration curve before
continuing to analyze samples.  If subsequent control standards also deviate by more than 10% from the
calibration curve, it is likely that the analyzer sensitivity has changed and a second calibration curve should
be generated. Sample analysis should continue after recalibration.
13. Calculation of Mercury Concentrations in Ambient Air

13.1 Calculation of Vapor Phase Mercury Concentrations

   13.1.1 Conversion of vapor phase mercury analysis results to ambient concentrations requires the results
from analysis (samples, blanks and calibration curve) and flow rate and duration data from field logs.  For
example calculations see Section 13.1.2.
      13.1.1.1 Vapor phase mercury concentrations in ambient air are reported in ng/m3.
      13.1.1.2 The amount of mercury collected on a Au-coated bead sample trap (ng Hg) is calculated from
the integrator response for the sample [in Peak Area Units (PAU)] multiplied by the slope of the calibration
curve, which is in ng Hg/PAU.
      13.1.1.3 The amount of mercury collected on a Au-coated bead sample trap is blank corrected by
subtracting the average amount of mercury collected  on field  blank traps. Field blanks are described in
Section 14.2.2.
      13.1.1.4 Vapor phase mercury samples are typically collected using 2 Au-coated bead sample traps
in series (the exit from the first trap is connected to the inlet of the second trap).  The total amount of Hg
collected for the sample is then calculated by simple addition of the blank corrected amounts of mercury from
the two sample traps.
      13.1.1.5 The volume (m3) of air sampled is calculated by multiplying the flow rate through the sample
traps (in cm3/min.) by the duration of the sample (min.) and converting the product from cm3 to m3.
      13.1.1.6 The ambient vapor phase mercury concentration in ng Hg/m3 is calculated from the total
blank corrected ng of Hg for the sample traps divided  by the cubic meters of air sampled.
   13.1.2  Example Calculation of Ambient Vapor Phase Mercury Concentration
      13.1.2.1 An example vapor phase mercury calibration  curve is given in Table 3 and discussed in
Section 12.5.2.
      13.1.2.2 Calculation of the ng of Hg recovered from two sample traps in series:

   Amount of Hg Recovered = Integrator Response x Slope of Calibration Curve

   7,135,900 PAU x  1.0585E-7ng Hg/PAU = 0.755ngHg
   236,183 PAU x 1.0585E-7ng Hg/PAU = 0.025ngHg

      13.1.2.3 Calculation of the total ng of Hg for the sample:

   Total Amount of Hg for sample = 2 (Sample, ng Hg - Average Field Blank, ng Hg)

   (0.755 ng Hg - 0.015 ng  Hg) + (0.025 ng Hg - 0.015 ng Hg)  = 0.750 ng Hg

      13.1.2.4 Calculation of the volume of air sampled at a flow rate of 0.3 Lpm and a sample duration of
24 hours, with average ambient temperature of 20°C and average barometric pressure of 750 mmHg:
June 1999              Compendium of Methods for Inorganic Air Pollutants            Page 5.0-25

-------
Method IO-5.0                                                                    Chapter IO-5
Vapor and Particulate Mercury	Mercury

   Volume of Air Sampled = Flow Rate x Duration

   (0.3 Lpm) (24 hr.) (1,440 min./24 hr.) (lQ-3m3/L) = 0.432 m3

      13.1.2.5 Correct total sample volume (m3) to standard conditions of 25°C and 760 mmHg:
                                   = (0.432
                                              ' '  760 A 301,
                                   = (0.432) (0.9868)  (0.9900)
                                   = 0.422 m3 at standard  temperature and pressure.
      13.1.2.6 Calculation of the vapor phase mercury concentration (ng Hg/m3) for an ambient air sample
at standard temperature and pressure:

   Concentration = Total Amount of Hg for Sample / Standard Volume of Air Sampled

   0.750 ng Hg / 0.422 m3 = 1.77 ng Hg/m3 at standard temperature and pressure.

13.2 Calculation of Particle-Phase Mercury Concentration

   13.2.1  Conversion of particle-phase mercury analysis results to ambient concentrations requires the
results from analysis (samples, blanks and calibration curve) and the volume of air sampled from field logs.
      13.2.1.1 Particle-phase mercury concentrations in ambient air are reported in pg/m3.
      13.2.1.2 The amount of mercury detected for the aliquot of sample analyzed is calculated from the
integrator response  [in Peak Area Units (PAU)] for the sample and the reagent blank. The difference between
the sample and reagent blank integrator response is multiplied by the slope of the calibration curve, which
isinpgHg/(PAU).
      13.2.1.3 The amount of mercury collected on the entire sample filter is calculated by multiplying the
amount of mercury  in the aliquot by the ratio of the total extraction volume of 20.5 mL (20 mL of extraction
solution and 0.5 mL of BrCl) to the volume of the aliquot analyzed (5 mL).
      13.2.1.4  The volume (m3) of air drawn through the sample filter is  calculated from the difference
between the "on" and "off" readings for the Dry Test Meter (DTM) used to measure sample volume and
adjusted by the calibration curve for the DTM display.
      13.2.1.5 The particle-phase mercury  concentration in pg/m3 is calculated from the amount of mercury
collected on the filter divided by the cubic meters of air sampled.
Page 5.0-26            Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5                                                                   Method IO-5.0
Mercury _ Vapor and Particulate Mercury

   13.2.2  Example Calculation of Ambient Particle-Phase Mercury Concentration
      13.2.2.1 An example particle-phase mercury calibration curve is displayed in Table 4 and discussed
in Section  12.6.7.
      13.2.2.2 Calculation of the pg of Hg detected for the aliquot analyzed:

   Amount of Hg Detected for Aliquot =
   (Sample - Reagent Blank Integrator Response) x (Slope of Calibration Curve)

   (7,135,900PAU- 168.320PAU) (9.1091E-5 pg Hg/PAU) = 635 pg Hg in aliquot.

      13.2.2.3 Calculation of the pg of Hg for the entire sample filter:

   Amount of Hg for Filter =  (Amount of Hg for Aliquot)  (Extraction Volume /Volume of Aliquot)

   (635 pg Hg) (20.5 mL) / (5  mL) = 2604 pg Hg for filter

      13.2.2.4  Calculation of the volume of air sampled from the DTM readings with average ambient
temperature of 20 °C and average barometric pressure of  750 mmHg during sampling:

   Volume of Air Sampled =  ("Off" - "On" DTM readings) x Slope of DTM Display Calibration
   (1075.6 m3 -  1031.3 m3) x 0.975 =  43.2 m3
      13.2.2.5 Correct total sample volume (m3) to standard conditions of 25 °C and 760 mmHg:

                                                     298
                                            760    273 + tA
                                              V760A301/
                                   =  (43.2) (0.9868) (0.9900)

                                   =  42.2  m3 at standard temperature and pressure.
      13.2.2.6 Calculation of the particle-phase mercury concentration (pg Hg/m3) for an ambient air sample
at standard temperature and pressure:

   Concentration =  [Amount of Hg for Filter - Avg. Field Blank]/ Volume of Air Sampled
   2604 pgHg/42.2m3 = 61.70 pg Hg/m3
14. Quality Assurance/Quality Control (QA/QC)

14.1 Personnel Qualifications
June 1999              Compendium of Methods for Inorganic Air Pollutants           Page 5.0-27

-------
Method IO-5.0                                                                      Chapter IO-5
Vapor and Particulate Mercury	Mercury
   14.1.1  Field operators should be properly trained in the methods for ambient air sample collection as
described in Section 11.
   14.1.2  Field  operators should routinely collect field blanks and shipping blanks to ensure that clean
handling techniques are consistently employed.
   14.1.3  Laboratory analysts should be properly trained in the procedures for analysis of ambient air
samples as described in Section 12.

14.2 QA/QC Samples

   14.2.1 Laboratory procedural blanks (LPB) are used to monitor the degree of background contamination
introduced during the laboratory analysis procedures. For analysis of vapor and particle-phase Hg this blank
is equivalent to the zero-point on the calibration curve and are analyzed before beginning analysis of samples.
   14.2.2  Field Blanks
      14.2.2.1 Field blanks are performed to determine the level of sample contamination during all phases
of sample handling, including:
   •  Sample collection and handling in the field
   •  Shipment
   •  Storage
   •  Sample handling and analysis in the laboratory
      14.2.2.2  Field blanks are  performed using  the same procedures as those described for collecting
samples in Section  11.  A minimum frequency of one field blank collected and analyzed per 10 samples is
recommended for each sampling media. Field blank levels less than 2% of the average amount of mercury
collected for a sample can be obtained using the methods described.
      14.2.2.3 Vapor phase mercury field blanks are performed by placing the filter pack and attached gold
bead trap in the sampling box as described for a sample, and left in the sample box for 2 minutes without the
vacuum line attached. The vapor phase mercury field blank is removed from the sampling box as described
for a sample and labeled appropriately.
      14.2.2.4  Particle-phase mercury field blanks are performed by loading a glass fiber filter  into the
open-face filter pack as described for a sample.  The filter pack is placed in the sampling box for 2  minutes
without connecting it to the vacuum line.  The particle-phase mercury field blank is removed from the filter
pack as described for a  sample and labeled appropriately.
   14.2.3  Collocated Samples
      14.2.3.1 Collocated samples are used to assess sample variability attributable to:
   •  Sample collection and handling in the field
   •  Shipment
   •  Storage
   •  Sample handling and analysis in the laboratory
      14.2.3.2  Collocated sampling for  a minimum  of 1  per 20 samples collected  is recommended to
properly evaluate variability.  Percent differences  of less than 10% between collocated samples can  be
achieved using the methods described.
      14.2.3.3  To collect collocated samples in the field, dual sets of sampling equipment must  be used
which are equivalent in all measurable ways.
   14.2.4  Storage Blanks
Page 5.0-28            Compendium of Methods for Inorganic Air Pollutants              June 1999

-------
Chapter IO-5                                                                     Method IO-5.0
Mercury	Vapor and Particulate Mercury

      14.2.4.1 Storage blanks are performed to determine the level of sample contamination during shipment
and storage of samples, and therefore are only required when supplies and samples are shipped between a
laboratory and a field site.
      14.2.4.2  A minimum frequency of one storage blank collected and analyzed per 10 samples is
recommended for each sampling media when samples are shipped to field sites.
      14.2.4.3 Storage blank levels less than 1% of the average amount of mercury collected for a sample
can be obtained using the methods described.
      14.2.4.4  For a vapor phase mercury storage blank, a gold trap is sent to the field site and back to the
laboratory for analysis without ever being opened in the field.
      14.2.4.5 Particle-phase mercury storage blanks are performed by loading a glass fiber filter from the
Teflon® storage jar directly into a plastic petri dish.  The packaged filter is then sent to the field site and back
to the laboratory for analysis, again without being opened.

14.3 Precision and Accuracy

   14.3.1 Precision
      14.3.1.1 Collocated samples can be used to assess the overall precision of the method (sampling and
analytical precision).  Precision of less than 15% can be achieved for ambient vapor phase and particle-phase
mercury measurements using the  methods described.
      14.3.1.2  For particle-phase mercury, samples analyzed in duplicate can be used to assess  analytical
precision. Analytical precision for the methods described should average less than  10%.
      14.3.1.3 For vapor phase mercury, samples can not be analyzed in duplicate.  Repeated injections of
vapor phase mercury standards can be used to assess the analytical precision which should average less than
5% for the methods described.
   14.3.2 Accuracy
      14.3.2.1 Accuracy can be assessed using standard reference materials (SRM) that have been analyzed
in a manner identical to the field  samples.  However, SRMs do not currently exist for ambient vapor and
particle  phase mercury.
      14.3.2.2  Comparison with other methods of analysis (i.e. neutron activation) or  inter-laboratory
comparisons can be used to  assess accuracy.
15. References

 1.  Bloom, N.S. and Fitzgerald, W.F., "Determination of Volatile Mercury Species at the Picogram Level
by Low-Temperature Gas Chromatography with Cold-Vapor Atomic Fluorescence Detection," Anal. Chim.
Acta., 208:151, 1988.

 2. Burke, J., Hoyer, M.,  Keeler, G., Scherbatskoy, T.,  "Wet Deposition of Mercury and Ambient Mercury
Concentrations at a Site in the Lake Champlain Basin,"  Water, Air, and Soil Pollution, 80:353-362, 1995.

 3.  Dumarey, R., Dams, R., and  Hoste, J., "Comparison of the Collection and Desorption Efficiency of
Activated Charcoal, Silver,  and Gold for the Determination of Vapor phase Atmospheric Mercury," Anal.
Chem., 57:2638-2643, 1985.
June 1999             Compendium of Methods for Inorganic Air Pollutants            Page 5.0-29

-------
Method IO-5.0                                                                  Chapter IO-5
Vapor and Particulate Mercury	Mercury

 4.  Dumarey, R., Temmerman,  E., Dams, R. and Hoste, J., "The Accuracy of the Vapor-Injection
Calibration of Mercury by Amalgamation/Cold-Vapour Atomic Absorption Spectrometry," Anal.  Chim.
Acta., 170:337-340,  1985.

 5.  Dvonch, J.T., Vette, A.F., Keeler, G.J., Evans, G., and Stevens, R.,  "An Intensive Multi-Site Pilot
Study Investigating Atmospheric Mercury in Broward County, Florida," Water, Air, and Soil Pollution,
80:169-178, 1995.

 6.  Fitzgerald, W.F., and Gill, G.A.,  "Sub-Nanogram Determination of Mercury by Two-Stage Gold
Amalgamation  and Gas Phase Detection Applied to Atmospheric Analysis," Anal. Chem., 15:1714, 1979.

 7.  Keeler, G., Glinsorn, G., and  Pirrone, N. "Particulate Mercury in the Atmosphere:  It's Significance,
Transport, Transformation and Sources" Water, Air, and Soil Pollution, 80:159-168, 1995.

 8.  Lamborg,  C., Hoyer,  M., Keeler,  G., Olmez,  I., and Huang,  X.  "Particle-Phase Mercury in the
Atmosphere: Collection/Analysis Method Development and Applications," inMercuryas a Global Pollutant:
Toward Integration and Synthesis,  Watras, C. and Huckabee, J. Eds., Lewis Publishers, 1994.

 9.  Bloom, N.S., "Determination  of Pricorgram Levels of Methylmercury  by Aqueous Phase Ethylation,
Followed by Cryogenic Gas Chromatography with CFAS" Can. J. Fish Aq. Sci., 46:1131, 1989.

10.  Bloom N.S., Prestbo E.M., Hall B.  and E.J. von der Geest, "Determination of Total Gaseous Hg in the
Ambient Atmosphere by Collection on lodated Carbon, Hot Acid Digestion and CFAS" Water,  Air and Soil
Pollut., 80:1315,  1995.

11.  Vermette S., Lindberg S., and N.S.  Bloom,  "Field Tests for a Regional mercury Deposition Network -
Sampling Design and Preliminary Test Results," Atmos. Envion., 29:11, 1995.

12.  Prestbo E. M., Liang L., Horvat M., and N.S. Bloom, Recent Advances in the Analytical Techniques for
the  Quantification of Mercury and  Mercury Compounds in Different Media,  USEPA 600-R-92-105 (1992).
Page 5.0-30            Compendium of Methods for Inorganic Air Pollutants             June 1999

-------
Chapter IO-5
Mercury
                                              Method IO-5.0
                                  Vapor and Particulate Mercury
          TABLE 1.  EXAMPLE OF AMOUNT OF Hg INJECTED FOR A TYPICAL
                       VAPOR PHASE CALIBRATION CURVE
Volume of Hg Saturated Air
Injected (,uL)
0
20
40
80
160
Amount of Hg Injected
(ng)
0
0.198
0.396
0.793
1.586
               Note:   Flask temperature = 16.6°C
                     Vapor density = 9.912 ng/cm3
                     1 cm3  = 1,000 ^L
              TABLE 2.  EXAMPLE OF AMOUNT OF Hg IN 5 mL ALIQUOT
                   ANALYZED FOR A TYPICAL PARTICLE-PHASE
                             CALIBRATION CURVE
Volume of 2 ng/mL Hg
Standard Added (,uL)
0
200
400
1,000
2,000
Amount of Hg in 5 mL
Aliquot Analyzed (pg)
0
97
191
465
889
June 1999
Compendium of Methods for Inorganic Air Pollutants
Page 5.0-31

-------
Method IO-5.0
Vapor and Particulate Mercury
                                                 Chapter IO-5
                                                     Mercury
             TABLE 3. EXAMPLE VAPOR PHASE Hg CALIBRATION CURVE
Hg Standard
Concentration,
PgHg
0
198
396
793
1,586
Response, Peak
Area Units (PAU)
41,193
1,866,300
3,729,482
7,451,226
15,083,592
Response-Zero
Point, Peak Area
Units (PAU)
0
1,825,107
3,688,289
7,410,033
15,042,399
Predicted Value,
PRHg
0
193
390
784
1,592
Percent
Difference from
Standard

-3
-2
-1
0
                            Slope = 1.0585E-7ngHg/PAU
                                   r2 = 0.9999
            TABLE 4.  EXAMPLE PARTICLE-PHASE Hg CALIBRATION CURVE
Hg Standard
Concentration,
pgHg
0
97
191
465
889
Response, Peak
Area Units (PAU)
168,320
1,144,900
2,158,900
5,204,200
9,993,400
Response-Zero
Point, Peak Area
Units (PAU)
0
976,580
1,990,580
5,035,880
9,825,080
Predicted Value,
pgHg
0
89
181
459
895
Percent
Difference from
Standard
0
-8
-5
-1
1
                            Slope = 9.1091E-5 pg Hg/PAU
                                   r2 = 0.9995
Page 5.0-32
Compendium of Methods for Inorganic Air Pollutants
June 1999

-------
Chapter IO-5
Mercury
                                                              Method IO-5.0
                                             Vapor and Particulate Mercury
   Secondary
   gold-bead
   trap

   Heated
   primary
   gold-bead
   trap
   Pre-filter
                              yapor-phase
                                  Hg
                                Collection
                                 System
                          Particle-phase
                               Hg
                            Collection
                             System
                                      0,3 Lpm
                                      low flow
                                       Inlet
                          N. 30 Lpm
                              high flow
                                inlet
                                                                         Polyuretliane tubing
                                                Black latex tubing
                                                Open face fitter pack for
                                                parttajlate Hg (glass fiber filter)

                                                Open face filter pack for metals
                                                (Teflon® filter)
                                                                       Variable transformer box
      Figure 1.  Example of ambient sampling system for collection of vapor and particle phase mercury.
June 1999
Compendium of Methods for Inorganic Air Pollutants
Page 5.0-33

-------
Method IO-5.0
Vapor and Particulate Mercury
                                                      Chapter IO-5
                                                          Mercury
                        Sampling Vapor and Particle Phase
                              Mercury in Ambient Air
 Project

 Site
 Location

 Date
 Height above ground

 Operator	
             Barometric Pressure (mmHg)

             Ambient Temperature (°C)

             Rain (Y/N)

             Relative Humidity (%)
                                         Before
After
                             CARTRIDGE INFORMATION
          Vapor-phase System
             Primary gold-bead "A" trap I.D. Number    	
             Secondary gold-bead "B" trap I.D. Number   	
             Heated primary trap temperature (°C)          	
          Particle-phase System
             Filter pack I.D. Number	
                            FIELD DATA INFORMATION
Clock
Time
(24-hr)





Flow
Check
(/)





Vapor-phase System
Mass Flow
Control
Setting





Flow Rate
(QJ
(Lpm)





Particle-phase System
Dry Gas
Reading





Flow Rate
(QJ
(Lpm)





Total
Sample
Time,
minute





Total
Sample
Volume,
L





Comments
    Figure 2. Field test data sheet (FTDS) for sampling vapor and particle phase mercury in ambient air.
Page 5.0-34
Compendium of Methods for Inorganic Air Pollutants
 June 1999

-------