SEPA
            United States
            Environmental Protection
            Agency
            National Training
            and Operational
            Technology Center
            Cincinnati OH 45268
            Water
Methods for the
Determination of
Chemical
Contaminants
EPA-430/1-80-006
April 1980
            Training Manual

-------
                                                   March 1980
                 PARTICIPANTS HANDBOOK

                          for

       METHODS FOR THE DETERMINATION OF CHEMICAL
             CONTAMINANTS IN DRINKING WATER
This Participants Handbook was developed by the Environ-
mental  Protection Agency, National Training and Operational
Technology Center with the Technical Support Division in
response to a request from the Office of Drinking  Hater.
 National  Training and Operational  Technology Center
          Office of Water Program Operations
         U.S.  Environmental  Protection Agency

-------
                     DISCLAIMER

Reference to conmercial products, trade names, or
manufacturers is for purposes of example and illustration.
Such references do not constitute endorsement by the
Office of Water Program Operations, U.S. Environmental
Protection Agency.

-------
  METHODS FOR THE DETERMINATION OF CHEMICAL CONTAMINANTS IN DRINKING WATER

                                INTRODUCTION


Course  Instructional Objectives:

This course  is designed to meet the needs for training those persons who will
be  involved  in analysis of potable waters for contaminants listed in the
Interim Primary Drinking Water Regulations.  The course can be used by either
State or Federal personnel.  Should the course be offered by a State, the
contents should be thoroughly checked to see if discrepancies exist between
the outlines prepared under the Federal regulations and what is required in
the State's  Drinking Water Act.

After successfully completing this course, the trainee should have sufficient
information  to carry out the various analyses.  The trainee will perform analysis
in  the  laboratory under the supervision of the offering authority's chemist.
The trainee  will be observed by the instructors and judged on the competence
of  laboratory technique and will be given a test upon completion of the course.

The course will cover the analytical methodology used to analyze for metals,
organics, nitrates and fluoride.  Additional topics covered will include
sample  handling and discussions on laboratory certification and the Act itself.

Although all parameters, inorganic and organic, are contained in this manual,
caution is advised in several areas.  First there is a provision in all acts,
both Federal and State, to allow alternate test procedures.  This provision can
be  carried out again on a Federal or State level.  Consequently, the users of
this manual  should inquire of their appropriate authority as to the existence
of  any additional methodology which has been approved.

Secondly, from time to time the Primary Drinking Water Regulations will be
amended and  new parameters which must be analyzed for added.  This will create
a need for constant revision of the manual on the part of the user to keep
it  current.  Another thought should be considered here and that is the Secondary
Regulations.  These will be issued on esthetic principles and will not be manda-
tory.  However, the methodology that should be used is listed in these regulations
and analysts might wish to examine these.

The exact make-up of the manual has been designed in such a way as to allow
maximum flexibility in determination of course content.  Options have been
provided so  as to allow the course director to select the final content.

For a brief  overview of the course, the content covered, and the schedule, turn
to  the Agenda.

A word about the handbook itself is in order.  The handbook is designed to be
used by you  from the time you receive it throughout the course, and when you
are back on  the job.  At the outset, there are some tasks the participant should
complete prior to attending the course.  During the course, the handbook will
be  referred  to daily or with each set of units and presentations.  It spells
out for you  what you will learn, how, when, under what conditions, and when
you will know if you have learned what is intended.   The instructors will
use this same material, with some variations depending on the situation,
participant  needs,  or scheduling requirements.

                                                                      A-l

-------
      The handbook is assembled in this loose-leaf form to allow the instructor some
      flexibility in unit selection or sequencing.  Also, it allows you to insert
      your notes from the daily sessions next to the corresponding unit.  Should
      the instructor or a fellow-participant provide additional information (a paper,
      articles, etc.) not envisioned for the course but relevant and useful, it can
      be inserted in the handbook.
A-2

-------
                                  CONTENTS

   METHODS FOR THE DETERMINATION OF CHEMICAL CONTAMINANTS IN DRINKING WATER

                             PARTICIPANTS HANDBOOK
Introduction                                                  A
Table of Contents                                             B
Pre-Course Activities                                         C
  Pre-Test                                                    Cl
  Biographic Statement                                        C2
  Participant Survey                                          C3
Course Agenda                                                 D
Instructional Units                                           E
  Introduction
  General  Topics
    Regulations Concerning the Safe Drinking Water Act        El
    Sample Handling and Preservation                          E2
    Basic  Statistics                                          E3
    Quality Control                                           E4
    Safety                                                    E5
  Inorganic Parameters
    Silver                                                    E6
    Extraction Techniques for Cd, Cr, and Pb                  E7
    Mercury                                                   E8
    Arsenic and Selenium                                      E9
    Nitrate
      Cadmium Reduction                                       E10.A
      Brucine                                                 E10.B
    Fluoride
      SPADNS                                                  Ell.A
      Electrode                                               Ell.B
      Distillation                                            Ell.C
    Barium                                                    El 2
    Chlorine and Turbidity                                    E13.A
    Measurement of Free Chlorine Utilizing the DPD Kit        E13.B
    Determination of Turbidity                                E13.C
Reserved for Future Expansion                           E14 - E19
  Organics
    Pesticides                                                E20
    Chlorophenoxys                                            E21
Elective/Optional Units                                       F
                                                                        B-l

-------
  METHODS FOR THE DETERMINATION OF CHEMICAL CONTAMINANTS IN DRINKING WATER

                          PRE-COURSE ACTIVITIES
Each of the following forms to be completed by you concern activities you are
asked to complete prior to the course itself.  Those are the Pre-test, Bio-
graphic Statement, and Participant Survey.  They are explained in detail  in
the introduction to each form.

A fourth sub-section is entitled Pre-Course Preparation.  It is simply those
materials or activities the instructor may appropriately deem important to
read or carry out prior to arriving for the course.   This part of the hand-
book is optional and for the instructor to include or exclude.

-------
                                             Course Pre-Test
Dear Participant:

As your first action in relation to this course, we are asking you to complete
a short pre-test and return it to us before coming to the course.   It is im-
portant to have this information in order to measure how effectively the course
is meeting its objectives.  This, plus the post-test given upon completing
the course, will help us with this measurement.

Note, we talk of "measuring the course," not measuring you in the  sense of
pass or fail.  While learning is the primary responsibility of the learner,
we recognize that course design, content, methodology, instructor, learning
environment, etc., all  play a role.  This measurement plus other feedback
the instructors will get, will assist us with future design and delivery.

You are encouraged to take this test without any outside help from books or
individuals.  This test is designed to answer only the question of have you
improved your knowledge by taking this course.   It will also give  some indica-
tion on course design and instructor teaching.   No grade will be assigned nor
is any other use intended.
                                                                        Cl-1

-------
                                  Pre-Test
1.  The analytical methods to be used in the analyses of Drinking Water Samples
    are set down in the

    a.  Interim Primary Drinking Water Regulations
    b.  Safe Drinking Water Act
    c.  Clean Water Act

2.  Organic samples are collected in                   containers.

    a.  glass or plastic
    b,  glass only
    c.  plastic only

3.  The maximum holding time recommended for metals is

    a.  14 days
    b.  180 days
    c.  30 days

4.  A statistical  measurement for precision is

    a,  percent recovery
    b.  central tendency
    c.  standard deviation

5.  Minimum quality control requires that daily checks of a standard curve
    be within             of the original curve.

    a.  + 10%
    b.  +  5%
    c.  j-  2%

6,  Safety practices should be carried out

    a.  at al1  times
    b.  only when  hazardous materials are being used
    c.  only when  the supervisors are watching

7.  The required analytical method, published in the Interim Primary Regulations,
    for silver is

    a.  dithizone
    b.  silver diethyldithiocarbamate
    c.  standard atomic absorption techniques

8.  The MCL for silver was based on                     considerations.

    a.  cost
    b.  aesthetic
    c.  health

-------
    9.  The oxidant and fuel  gases used when silver is determined are

        a.  air-acetylene
        b.  nitrous oxide-acetylene
        c.  argon-hydrogen

   10.  In order to determine cadmium, chromium and lead at their MCL's, the
        sample needs to

        a.  be solubilized
        b.  extracted and concentrated
        c.  both

   11.  The oxidant and fuel  gases used to determine cadmium, chromium and lead
        are

        a.  air-acetylene
        b.  nitrous oxide-acetylene
        c.  argon-hydrogen

   12.  The extraction technique

        a.  necessitates doing each metal (cadmium, chromium, lead) separately
        b.  allows all metals (cadmium, chromium, lead) to be done with one extraction
        c.  need a preliminary colorimetric procedure.

   13.  The determination of mercury  is carried out by

        a.  normal atomic absorption  techniques
        b.  colorimetricly
        c.  a flameless atomic absorption technique

   14.  The oxidant and fuel  gases used to determine mercury

        a.  are air-acetylene
        b.  nitrous oxide-acetylene
        c.  air only

   15.  The organic forms of mercury  are                to convert to metallic form.

        a.  difficult
        b.  easy
        c.  impossible

   16.  The oxidant and fuel  gases used to determine arsenic and selenium
        are

        a.  air-acetylene
        b.  nitrous oxide-acetylene
        c.  argon-hydrogen

   17.  Organic forms  of arsenic  are  analyzed               by the gaseous hydroxide
        method.

        a.  directly
        b.  after  an oxidation step
        c.  colorimetricly
Cl-4

-------
18,  The zinc slurry provides                in the gaseous hydride procedures
     for arsenic and selenium.

     a.  for the reduction
     b.  the hydrogen for the flame
     c.  the hydrogen to form the hydride

19,  Nitrate is determined               in the cadmium reduction method.

     a.  as nitrate
     b.  as nitrite
     c,  as cadmium

20.  Nitrate samples for the reduction column should not be preserved
     with

     a.  sulfuric acid
     b.  refrigeration at 4°C
     c.  mercuric chloride

21,  The nitrate sample for the cadmium reduction method is filtered to remove
     turbidity which could

     a.  react with the nitrate
     b.  oxidize the nitrate to nitrite
     c.  restrict flow through the column

22.  The brucine test analyzes nitrate as

     a.  nitrate
     b.  nitrite
     c.  brucine

23.  One extremely important control in the brucine test is

     a.  size of the particles
     b.  concentration of nitrite
     c.  temperature

24.  The brucine-nitrate test is a                  test.

     a.  colorimetric
     b.  atomic abosrption
     c.  titrametric

25.  For drinking water samples                must precede the SPADNS test.

     a.  filtration
     b.  use of the electrode
     c.  distillation

26.  Fluoride samples are preserved by the addition of

     a.  nothing
     b.  nitric acid
     c.  mercuric chloride


                                                                        Cl-5

-------
   27.  The SPADNS method for fluoride  is a                      procedure.

        a.  colorimetric
        b.  atomic absorption
        c.  titrametric

   28.  The distillation procedure works by

        a.  distilling over  the  interferences  and  leaving  the  F  behind
        b.  distilling over  the  F" and  leaving the interferences behind
        c.  forming  a color  with the  interferences

   29.  A  new  batch  of acid/water mix must be  used

        a.  with  each sample
        b.  after three samples
        c.  when  the solution turns brown

   30.  If the temperature  is allowed to go  beyond 180°C

        a.  the fluoride  is  not  distilled
        b.  the iron carries over
        c.  sulfate  is carried over

   31.  For drinking water  samples             must precede the electrode method.

        a.  nothing
        b.  distillation
        c.  filtration

   32.  The electrode must  be connected to                for the  F~ determination.

        a.  pH meter with expanded  scales
        b.  specific ion  meter
        c.  either of the above

   33.  The electrode itself and the                account for the small number
        of interferences.

        a.  distillation
        b.   the TISAB buffer
        c.  the complexone

    34.  Barium is determined by

        a.   colorimetry
         b.  atomic absorption
         c.  titrametry

    35.  The barium samples  are  preserved by  adding

         a.  nitric acid
         b.  sulfuric acid
         c. mercuric chloride
Cl-6

-------
36,  In order to express the value of'total" barium, a              step
     must be performed.

     a.  filtration
     b.  weighing
     c.  solubilization or digestion

37.  The approved method for residual chlorine determination for water supply
     samples is the

     a.  o-tolidine
     b.  phenol red
     c.  DPD

38.  The kit form of the approved method

     a.  can be used
     b.  cannot be used
     c.  must be applied for under alternate test procedures

39.  Chlorine samples

     a.  can be preserved overnight
     b.  can be held for 48 hours
     c.  cannot be preserved

40.  The turbidity sample must be taken

     a.  in the plant
     b.  at an entry point to the distribution system
     c.  in the distribution system

41.  The reason for the MCL on turbidity is because

     a.  it may interfere with disinfection
     b.  it makes water look bad
     c.  it makes water taste bad

42.  Turbidity measurement must be carried out

     a,  a number of times based on population served
     b.  once a week
     c.  once a day

43.  The Interim Primary Drinking Water Regulations

     a.  become effective in December of 1977
     b.  became effective in December of 1975
     c.  became effective in June of 1977

44.  The Interim Primary Drinking Water Regulations include maximum contaminant
     levels (MCL's) as well as monitoring frequencies for

     a.  chemical, bacteriological, radiological  contaminants
     b.  chemical, bacteriological contaminants
     c.  chemical  contaminants


                                                                        Cl-7

-------
   45.  The maximum holding time for the chlorinated hydrocarbons samples is

        a.  14 days
        b.   7 days
        c.  none

   46.  The pesticides which are to be monitored are

        a.  endrin, aldrin, lindane, methoxychlor, toxaphene
        b.  endrin, lindane, methoxychlor, aldrin
        c.  toxaphene, methoxychlor, lindane, endrin

   47.  The pesticides are extracted from the sample using

        a,  hexane
        b.  petroleum ether-ethylether
        c.  hexane-methylene chloride

   48.  The chlorphenoxy  herbicides to be monitored for are

        a.  2,4,  D; 2,4,5 TP;  2,4,5 T
        b.  2,4,  D; 2,4,5 T
        c.  2,4,  D; 2,4,  5 TP

   49.  The herbicides are extracted from the sample using

        a.  hexane
        b.  hexane-methylene chloride
        c,  ethylether

   50.  The maximum holding time for herbicide  samples  is

        a.  14 days
        b,   7 days
        c.  none
Cl-
   P

-------
                            BIOGRAPHIC STATEMENT


Dear Participant:

We ask that you answer and complete the following items as briefly as
possible.  The purpose is simple:   to acquaint the instructor with his/
her students prior to the course.


1)  Name:  (please print)                              2)   Date:

3)  Address:


                                                       4)   Date of Birth;

5)  Present Position or Job Title:
6)  Major job functions or responsibilities:
7)  Courses or study undertaken in relation  to job:
                                                                        C2-1

-------
             PARTICIPANT SURVEY FOR THE METHODS OF ANALYSIS FOR
                 INORGANIC CONTAMINANTS OF POTABLE WATERS
1.  Identify the experience you have with each method for the various
    contaminants covered in the course (give approximate length of
    experience).
2.  How much experience do you have in atomic absorption?
3.  Do you have sufficient basic laboratory skills to carry out necessary
    laboratory procedures (use student skills checklist)?
4.  List the goals which you wish to achieve in attending this course.
5.  Will there be a need for you to pass on the information attained in this
    course to others?
                                                                         C3-1

-------
                                            Name	
                                            Employer 	

                          STUDENT SKILLS CHECKLIST
To assist us in processing applications, please check YES or NO for each of the
following items:
                                                                     YES     NO
I have operated a laboratory gas burner		   	
I have operated a laboratory hotplate/stirrer	___   	
I have operated an autoclave		   	
I have operated a laboratory drying oven	  	   	
I have used a vacuum source to filter liquids		
I have used a desiccator	_      	
I have weighed items on an analytical balance		
I have weighed items on a double pan balance  .....  	  	   	
I have used a graduate to measure liquids	  __   	
I have used a volumetric pipet to measure liquids		   	
I have used a graduated pipet to measure liquids	  .	   	
I have used a pipet bulb to fill a pip^t		
I have used mouth suction to fill a pipet		
I have used an inoculating loop to transfer small amounts of liquid  	   ___
I have used disinfectant to sterilize a lab bench work area ....  	   	
I have poured liquid from a container into glass test tubes ....  	   __
I have prepared media used for coliform tests 	  	
I have used chromic acid to clean glassware	  .   	   	
I have operated a laboratory safety shower  	
I have operated a laboratory eye washer 	  	
I have operated a fume hood	  	
I have prepared manganous sulfate solution  	  	
I have made out labels for bottles  or reagents		   	
I have used a buret		   _
I have used starch as a chemical change indicator		    	
I have titrated one solution against another to  a color  change  end
  Point 	
I have recorded a reading at a meniscus 	
I have recorded laboratory data in a laboratory notebook  	
I have entered laboratory data on a pre-printed form  .......
I have recorded information about samples  on record sheets   .  .  .  .
I have located required purchase information in a  catalog of
  laboratory equipment  	
I have written a  purchase order for chemicals  to be used in  the  lab

-------
                                                 Name 	
                                                 Employer
                                                                         YES    NO
     Volume means space occupied by a solid, liquid, or gas .  . . .
     mg/1 means milligrams per liter  .... 	
     Normality (N) is a way to express concentration in a solution
     1 kilogram equals 0.001 gram 	
     1 inch equals 2.54 cm  	
     1000 ml equals 1 liter	 .  . . ,
     85 times 4.1 equals 42.5	
     7 minus 2 divided by 0.02 equals 250 	
     3.26 rounded to the nearest tenth is 32.6  	
     84.55147 rounded to the nearest thousandth is 84.551 . .  . . ,
C3-4

-------
                            INSTRUCTIONAL UNITS

                                INTRODUCTION


The material covered in this section represents the core of the course.  All the
parameters listed in the Interim Primary Regulations are covered.  Where more
than one analytical procedure has been permitted, all are included here.  The
analyst will have to make the choice as to which will be used.  The material has
been written in a format that allows the trainee to utilize the outline as the
analysis is carried out.  The step-by-step outline proceeds in the same order as
an analyst must proceed.  Each step is given and if warranted, additional informa-
tion is given in the next column.  The student should be cautioned to read each
step and the material, if any, in the information column before doing the step.

Also included in the outline is an equipment list which lists capital, reuseable
and consumable items necessary to perform the analysis.  Where more than one
analysis is being performed, the equipment lists will have duplication and the
trainee must purchase only what is needed.

The level of training required to perform each analysis varies.  For example, the
outlines on residual chlorine and turbidity need little formal chemical educational
background to be performed.  However, the atomic absorption and gas chromatographic
analysis require considerable background to perform.  The later two analyses should
be performed only by experienced chemists or done under their supervision.

As new persons enter into the laboratory and assume responsibility for certain
analyses, this manual or sections of it can be used to acquaint or refresh the
analyst with the methodology for which he is responsible.
                                                                         E-l

-------
                REGULATIONS  CONNECTED WITH THE SAFE DRINKING WATER ACT

    I.   Introduction

        The  primary interest of  the Safe Drinking Water Act was to produce potable
        water  for  the  consuming  public.  It  is estimated that there are 240,000
        water  supplies  in  the United States.  That is about 40,000 community supplies
        and  about  200,000  non-community water supplies.  The Safe Drinking Water Act
        required that  limits be  set down for those materials which, when found in
        natural  waters, would pose a health  hazard.  From this requirement a document
        was  drawn  up and published for public comment.  This was the Proposed
        Interim  Primary Regulations.  After  the comments were received and acted
        upon,  the  Interim  Primary Regulations were published in the Federal Register.
        These  Interim  Primary Regulations listed the health related agents to be
        monitored  for  in the nations drinking waters.  A monitoring frequency was
        set  down for the community and non-community supplies and a "Maximum
        Contaminant Level" (MCL) for each was set.  In addition, a list of "approved"
        methods  for the analysis of each parameter was given.

        The  "Act"  stipulated that a study was to be made by the National Academy
        of Science (NAS) on  the  parameters that were included in the Interim
        Primary  Regulations.  This study was to look at other compounds to in-
        clude, if  some parameters should be excluded, if the MCL were right and
        suggest  changes and  areas of research.  After the NAS report was finished
        the  Interim Primary  Regulations were to be revised and published for public
        comment  and then promulgated.

        The  Environmental  Protection Agency was to publish a list of parameters and
        methods  to analyze for these parameters which a treatment facility might
        wish to  monitor for.  These parameters were to be based on aesthetic values
        and  were not to be of an enforceable nature.  These "Secondary Regulations"
        were to  follow the route of the Primary in that they were to be proposed
        and  sent out for public comment.

        In addition the the  Primary regulations, EPA was called on to issue any
        additional  parameter it felt might endanger the health of the consuming
        public.  These additional parameters after being commented on would be
        a part of  the already published Primary Regulations.  As with the other
        parameters analytical methodology and monitoring frequencies would be
        published  for the  new parameters.

        After  publication  of the analytical  methodology in the Primary Regulations,
        any  new  methods could be used after being ajudged comparable to those already
        published.   A review panel  and an approved mechanism was to be set up
        to supervise the requests for "Alternate Test Procedure Approval."  If the
        new  method was found to be comparable, it could be approved on a national
        or regional area level.

  II.   National Interim Primary Drinking Water Regulations

        Section  1412 of the Safe Drinking Water Act requires the Environmental
        Protection Agency  to publish proposed national  interim primary drinking
        water  regulations.   These regulations were to be proposed 90 days  after
        the enactment of the Safe Drinking Water Act.   Public comments were to be
        solicited on the proposed Interim Primary Drinking Water Regulation and
        these  regulations were to be promulgated 180 days after enactment.

AT.WMP.4.3.80
                                                                               Fl-1

-------
     The  Proposed  Interim  Regulations were  issued on March  14,  1975, and promul-
     gated  on  December  25,  1975.  Since the Act states that these  regulations
     should become effective  18 months after promulgation,  they became effective
     June 24,  1977.   These  Interim  Primary  Regulations will be  revised on an as
     needed basis  as  well  as  at least every three years.

     Since  the Interim  Primary Regulations  are based on health  hazards to the
     consuming public,  these  regulations  contain a  list of  Organic and Inorganic
     materials and a  maximum  contaminant  level for  each.  These levels are based
     upon possible health  hazards that may  occur after a  lifetime  of consuming
     approximately two  liters of water per  day.  The regulations set down what
     must be done  by  a  water  supply should  one of these maximum contaminant
     levels (MCL)  be  surpassed.  Also, a  part of the regulations is a list of
     approved  methods for  the analysis of the parameters  listed.   The parameters,
     their  levels  and methods are listed  in Table 1.

     Also included in the  regulations are the sampling and  monitoring frequencies
     for  the published  parameters.   This  sets down  how often  samples must be
     analyzed  and  for which of the  parameters.  Microbiological parameters are
     also listed  in the Interim Regulations.  The limits  and  sampling frequencies
     are  covered  in other  manuals.   The chemical sampling and monitoring require-
     ments  are listed in Table 3.

     Under  the microbiological sections in  the  Interim Primary  Regulations a
     supply may substitute residual  chlorine determinations for a  portion of
     the  microbiological tests.  The level  of free  chlorine that must be main-
     tained is 0.2 mg/1.   The analytical  method to  determine  this  is the DPD
     colorimetric  or  titrimetric method.   It was the intent to  allow the use
     of the color  comparator  kits for this  analysis.  Due to  the inability
     to preserve  a free chlorine sample it  is expected that the operator him-
     self will perform  this test.

     On July 19,  1979,  a proposed amendment to  the  National Interim Primary
     Drinking  Water Regulations was published.  As  with any proposed rule it
     calls  for public comment after which they will  be promulgated and become
     an addition  to the NIPDWR.

     The  analytically significant areas in  these proposed regulations are the
     publication  of alternative analytical  techniques approved  for nationwide
     use.  These  are  shown in Table 5.  In  addition, the  community supplies
     are  required  to  monitor  for sodium,  at least annually  for  systems utilizing
     surface water sources and at  least every three years for systems solely
     utilizing ground waters  sources.  Analyses for sodium  are  to  be carried
     out  by either flame photometric or atomic  absorption methods, the references
     for  which are given in Table  5.

     Some supplies when so notified by the  State will be  required  to  initiate
     a corrosion  control program.   This is  designed to protect  the drinking
     water  from possible corrosion  products as  lead, cadmuim, asbestos and
     organic compounds. The  proposed  rules suggest three ways  to  calculate
     some form of corrosion index  and  a suggested  limit for each method  of
     calculation.  When these proposed rules  are commented  on,  it  is hoped
     some decision can  be  made on  this point.
El-2

-------
      Some changes connected with non-community supplies only are suggested in
      the proposed armiendments.  These would extend for an additional  year the
      nitrate monitoring deadline and allow up to 20 mg/1  of nitrate in some
      systems at the discretion of the State.

      In addition to the proposed amendments, a final  regulation concerning the
      Control of Trihalomethanes in Drinking Water were published on November
      29, 1979.  These regulations set an MCL of 0.1 mg/1  of total  trihalomethane
      (TTHM).  This is the arithmetic sum of the concentrations of chloroform,
      dibromochloromethane, bromodichloromethane and tribromomethane,  rounded
      to two significant figures.

      These regulations would effect, at the present time, only those  supplies
      serving populations greater than 10,000 persons.   There is a  phase in
      approach for both monitoring and an effective date for the MCL for supplies
      of different sizes.  The initial monitoring frequency would be four samples
      per quarter taken all on the same day at different locations  in  the distri-
      bution systems.   The analytical methods that may  be  used are  the purge and
      trap technique or the liquid-liquid extraction technique, both of which
      are published in the Federal  Register along with  the regulations.   In
      addition, the following references are given.

           1)  "The Analysis of Trihalomethanes in Finished Waters  by  the
                Purge  and Trap Method," Method 501.1 EMSL, EPA, Cincinnati,
                Ohio  45268.

           2)  "The Analysis of Trihalomethanes in Finished Waters  by  the
                Liquid/Liquid Extraction Method," Method 501.2, EMSL,  EPA,
                Cincinnati, Ohio  45268.

III.   National  Academy of Science  Study

      After publication of the Proposed and Interim Primary Regulations  the EPA
      is required by the Act to enter into arrangements with the National
      Academy of Sciences to conduct  a study.  This study  should determine the
      maximum contaminants levels  which should be recommended under the  revised
      primary regulations in order to protect the health of persons from any
      known or anticipated adverse effects, and the existence of any contaminants -
      the levels of which in drinking water cannot be determined but which may
      have an adverse  effect on the health of persons.

      The study was to be presented to Congress no later than 2 years  after the
      date of enactment of the Act.   EPA will use this  study in deciding whether
      to include any such contaminants in a revised Primary Drinking Water
      Regulations.

      In conducting its study the  National Academy of Sciences is directed to
      consider only what is required  to protect public  health, not  what  is
      technologically  or economically feasible or reasonable.   Based on  the results
      of the NAS study, EPA may specify additional  contaminants with adverse
      health effects.   It may establish new maximum contaminants levels.   It may
      prescribe a list of known water treatment techniques which will  reduce the
      concentration of any contaminant for which no maximum contaminant  level  is
      established (e.g., viruses,  organics, asbestos),  or  it may establish
      requirements  for operation and  maintenance.
                                                                               El-3

-------
                                  TABLE  1
Parameter
Inorganic -
Arsenic
Barium
Cadmium
Chromium
Lead
Mercury
Nitrate
Selenium
Silver
Turbidity
Fluoride

Limit
mg/1 iter
Other than Fl
0.05
1.00
0.01
0.05
0,05
0.002
10.00
0.01
0.05
1.0 TU
Temp. Dep.
See Table 2

Method
uoride
Atomic Absorption
(gaseous hydride)
Atomic Absorption
(Std, Conditions)
Atomic Absorption
(Std. Conditions)
Atomic Absorption
with Chelation Ext.
Atomic Absorption
with Chelation Ext.
Flameless Atomic
Absorption
Brucine (Co 1 or i metric)
Cadmium Reduction
(Colorimetric)
Atomic Absorption
(gaseous hydride)
Atomic Absorption
(Std. Conditions)
Nephelometric
Electrode
SPADNS (Colorimetric)
with distillation

EPA1
95-96
97-98
101-103
105-106
112-113
118-126
201-206
145
146
295-298
65-67
59-60
Reference
Std. Meth. (13th) ASTM5

210-215
210-215
210-215
210-215
396-402
461-464

210-215
350-353
172-174 342-344
171-172 340-342
174-176
El-4

-------
                           TABLE  I  (Cont'd.)
Parameter
Organic
Chlorinated
Hydrocarbons
Endrin
Lindane
Methoxychlor
Toxaphene
Chlorophenoxys
2.4-D
Si 1 vex
Limit
mg/ liter



0.0002
0.004
0.1
0.005

0.1
0.01
Method


Gas Chroma tography




Gas Chroma tography


Reference
ASTM
*3
EPA Method J 609-624




EPA Method 4 595-602


1
2
Methods for Chemical  Analysis  of Water and  Wastes,  EPA,  Office  of  Technology
Transfer, Cincinnati, Ohio 45268.

Standard Methods for  the Examination  of Water and Wastewater, 13th  Edition,
1971, APHA, 1015 18th St., NU, Washington,  DC 20036,
 Method for Organochlorine  Pesticides  in  Industrial  Effluents, EMSL, EPA,
 Cincinnati, Ohio 45268.

 Method for Chlorinated Phenoxy  Acid Herbicides  in  Industrial Effluents, EMSL,
 EPA,  Cincinnati, Ohio  45268.

 Annual Book of ASTM Standards,  1977.  Part 31.
           Temperature  °C

           12.0 and  below
           12.1  to 14.6
           14.7 to 17.6
           17.7 to 21.4
           21.5 to 26.2
           26.2 to 32.5
                                   TABLE 2

                                Fluoride MCL
                                                  MCL mg/1

                                                     2.4
                                                     2.2
                                                     2.0
                                                     1.8
                                                     1.6
                                                     1.4

-------
                                           TABLE 3
          Type of
          Contaminant
          Inorganic
          Organic
          Turbidity
Water
Source
Surface

Ground


Surface


Ground


Surface
   Public Water Supply
  Sampling Requirements

    Sampling Period
Ini tated by^ Conclude By

Community Water Supplies

 6/24/77     6/24/78
                                   6/24/77
             6/24/79
 6/24/77     6/24/78
 To be specified by
      State

 6/24/77
Sampling Frequency
    Thereafter
at yearly intervals

every three years
If MCL is
Exceeded
collect 3
additional
samples in
one month
every three years or
as required by State

To be specified by State
must measure daily
resample
within one
hour
          Inorganic-
          Nitrate only
          Turbidity
       Non-Community Water Supplies

Surface   6/24/77     6/24/79
  and
Ground
Surface   12/24/77
                          To be specified by
                          State
                          must be measured
                          daily
                       collect 3
                       additional
                       samples in
                       one month

                       resample
                       within one
                       hour
          Report to State any violations within 48 hours,  including violation  of  monitoring
          requirements.

          Report to State tests results  within 40 days.
El-6

-------
IV.   National Secondary Drinking Water Regulations

     The Safe Drinking Water Act,  Section  1412  C,  required  the  EPA  to  propose
     national secondary drinking water regulations.   These  regulations  were  to
     be propsoed within 270 days after the enactment of the Safe  Drinking  Water
     Act.  Again, public comment was  to be requested and acted  upon before
     promulgation of the secondary regulations.  These  secondary  regulations
     are to be based on contaminants  that  tend  to  make  water disagreeable  to
     use, but that do not have  any particular adverse public health effect.

     The secondary regulations  were proposed on  March 31, 1977, and appeared  in
     final  form on July 19, 1979,  and included  the following:
                                    TABLE  4
     Contaminant

     Chloride

     Color

     Copper

     Corrosivity

     Foaming Agent

     Iron

     Manganese

     Odor
   Li mi t

250 mg/1

15 Color Units

1.0 mg/1

Non-Corrosi ve

0.5 mg/1

0.3 mg/1

0.05 mg/1
      Method
     Reference
EPA    14th Std,  Meth,
Potentiometric

Platinum-Cobalt

Atomic Absorption

Reserved

Methylene Blue

Atomic Absorption

Atomic Absorption
3 Threshold Odor Consistent Series
Number
     Sulfate

     TDS

     Zinc
6.5 - 8.5

250 mg/1

500 rng/1

5 mg/1
Glass Electrode

Turbidimetric

Total Residue

Atomic Absorption

36-38
108-109
157-158
110-111
116-117
287-294
239-240
277-278
270-271
155-156
306
64
144
600
144
144
75
460-465
496
91
144
    These secondary regulations are applicable to all public water systems, but
    are not enforceable on the Federal level and are intended as guidelines for
    the states.  However, depending on their legislation, may be enforceable
    by the State.
                                                                           El-7

-------
       V.  Alternate  Test  Procedures Approval
           Section  141.27  of  the  National  Interim Primary Drinking Water Regulations
           permitted  the establishment of  a  procedure for Approval of Alternate
           Analytical Methods.  On March 10,  1977,  this procedure was established.
                                       are  allowed, a case by case procedure and a
                                             Alternative analytical techniques
                                            be  published as amendments to the
Two parallel  approval  chains
procedure for national approval.
approved for nationwide yse will
           Interim  Primary  Regulations  in  the  Federal Register.

           Several  alternate methods  have  been approved  for  nationwide use and have
           been  published  in the  Federal Register.
                                       TABLE  5
           Measurement

           Arsenic


           Arsenic
           Barium


           Cadmium


           Chromium


           Fluoride



           Fluoride
           Fluoride
           Fluoride
                                    Method
                   1
                   1
 Flameless  Atomic Absorption,  Graphite  Furnace
 Technique.  Method  206.2.

 Silver  Diethyldithiocarbamate  Method,  Ref:
 "Methods for  Chemical Analysis  of Water  and
 Wastes," pp.  9-10,  EPA,  Office of Technology
 Transfer,  1974.  Method  206.4.

 Flameless  Atomic Absorption,  Graphite  Furnace
 Technique.

 Flameless  Atomic Absorption,  Graphite  Furnace
 Technique.

 Flameless  Atomic Absorption,  Graphite  Furnace
 Technique.

 Automated  Alizarin  Fluoride  Blue, Ref:   "Standard
 Methods for the  Examination  of Water and Wastewater,"
 14th,  pp.  614-616,  1975.
?
"Zirconium-Eriochrome Cyanine R, Ref:   "Methods  for
 Collection and Analysis  of Water Samples for
 Dissolved  Minerals  and Gases," USGS, Book 5, Chapter
 A 1,  pp. 90-93.

 Modified Automated  Alizarin  Fluoride Blue.  Ref.
 "Fluoride  in  Water  and Wastewater Industrial Method
 #129-71W"  December  1972,  Technicon  Industrial  Systems,
 Tarrytown,  Hew York 10591.

 Automated  Electrode Method.   Ref:   "Fluoride in Water
 and Wastewater," Technicon Industrial  Method #380-75WE."
 February 2, 1976.   Industrial  Systems,  Tarrytown,
 New York 10591.
El-8

-------
Measurement

Lead


Mercury




Nitrate
                 Method
1
Nitrate



Organics



Organics (Pesticides)



Organics (Herbicides)



Selenium


Selenium



Silver


Turbidity


Sod ium
1
 Flameless Atomic Absorption, Graphite Furnace
 Technique.

 Automated Cold Vapor Technique,  Ref:   "Methods
 for Chemical  Analysis of Water and Wastes,"
 pp. 127-133,  EPA,  Office of Technology Transfer,
 1974.

 Automated Hydrazine Reduction, Ref:   "Methods for
 Chemical  Analysis  of Water and Wastes," pp.  185-
 194, MERC,  Analytical Quality Control  Laboratory,
 1971.

 Automated Cadmium  Reduction, Ref:   "Methods  for
 Chemical  Analysis  of Water and Wastes,"  pp.  207-
 212, EPA, Office of Technology Transfer,  1974.
?
"Gas Chromatographic, Ref:   "Methods  for Analysis  of
 Organic  Substances in Water," USGS,  Book  5,  Chapter
 A3, pp.  24-39.

 "Standard Methods  for the Examination  of  Mater  and
 Wastewater."   14th ed.  1975.  Qrganochlorine   Pesti-
 cides,  Part 509A,  pp. 555-564.

 "Standard Methods  for the Examination  of  Water
 and Wastewater."  14th  ed.  1975.   Chlorinated
 Phenoxy  Acid  Herbicides,  part 509B,  pp.  565-569.
?
"Hydride  generation - atomic absorption spectro-
 photometry, USGS,  Method,  1-1667-77,  1976.

 Flameless Atomic Absorption, Graphite  Furnace
 Technique,  Ref:  Atomic  Absorption Newsletter.
 11, No.  5,  pp. 100-116,  1975.

 Flameless Atomic Absorption, Graphite  Furnace
 Technique.
j
 Nephelometric method with styrene  Divinylbenzene
 Polymer  Standards.

 "Standard Methods  for the Examination  of  Water  and
 Wastewater,"  14th  ed.,  pp.  250-253 or  "Methods  for
 Analysis  of Water  and Wastes,  p.  147.
 The various furnace devices are considered to be atomic  absorption techniques-
 Methods of standard addition are to be followed as noted  on p.  78 of "Methods
 for Chemical  Analysis of Water and Wastes," EPA, Office  of Technology  Trans-
 fer, 1974.
                                                                       E1-9

-------
      2
       Copies  available from:   Water  Quality  Branch,  National  Center,  U.S.
       Geological  Survey,  112201  Sunrise  Valley  Drive,  Reston,  VA  22092.
      3
       Only the six pesticides  named  in the Interim  Primary  Drinking  Water
       Regulations are included:   Endrin,  Lindane, Methoxychlor, Toxaphene;
       2,4-D;  and  2,4,5-TP (Silvex).   Federal  Register,  Vol.  40, No.  248,
       pp.  59570-59571, December  24,  1975.
      4
       Additional  information  on  this method  is  available  from the Environmental
       Monitoring  and Support  Laboratory.   Commercial  products  of  Amco-AEPA-1
       Polymer are available from AMCO Standards International,  Inc.,  230 Polaris
       Ave.,  No.  C, Mountain View, California  94043.
El-10

-------
VI.  Certification

     Section 1401  (1)  of the Safe Drinking  Water  Act  defines  "Primary  Drinking
     Water Regulations"  to  include "quality control and  testing  procedures" to
     insure compliance with  maximum contaminant levels.   Pursuant  to the Act,
     the National  Interim Primary Drinking  Water  Regulations, Section  141 and
     142 require  that  for compliance purposes, "samples"  will be considered
     only if they have been  analyzed by  a laboratory  approved by the State,
     except that  measurement for  turbidity  and free chlorine  residual  may be
     performed by any  person acceptable  to  the State, and the State must
     establish and maintain  a program for certification  of laboratories
     conducting measurement  of drinking  water contaminants.

     A "Manual  for the Interim Certification of Laboratories  Involved  in
     Analyzing Public  Drinking Water Supplies" has been  compiled.  This manual
     describes  how the Environmental  Protection Agency will carry  out  a tentative
     program for  interim approval  and certification of its  ten Regional labora-
     tories and principal  State laboratories.  States without certification
     programs  are encouraged to use this program  as a model; States with equi-
     valent or better  certification programs are  encouraged to continue and
     improve.

     The manual describes evaluation procedures and minimum technical  requirements
     recommended  for certifying laboratories analyzing public drinking water
     supplies.  In addition  to identifying  requirements  that are critical to
     generation of valid data, optional  certification requirements have been
     included  as  guidance.
                                                                            El- 11

-------
                                    SAMPLING

   I.   INTRODUCTION

       With the intent of the Safe Drinking Water Act being the insurance of
       proper drinking water quality,  meaningful  analysis  of the water is im-
       perative to know if the water meets  the standards.   This analysis  can
       only be meaningful  if it is performed on a samole that is representative
       of the water to be analyzed.  Consequently,  the proper sampling technique,
       use of proper containers,  proper preservation  and adherence to  the set
       frequency of sampling must be carefully observed.

       In many instances  the laboratories  themselves  will  not be responsible
       for sampling.   However,  it is necessary that all  laboratories be aware  of
       what constitutes  a  representative,  properly  taken sample.   It  is the  re-
       sponsibility of all  laboratories sampling  for  parameters under  the Safe
       Drinking Water Act  to call  for  a resample  if the  sample does not meet proper
       sampling procedures.   To analyze a  sample  which has  been doubtfully taken
       is to present data  which is dubious  in meaning.   If  the laboratory is
       responsible for taking the samples,  it is  doubly  important that the persons
       in the laboratory  be  aware of proper techniques.

       There is at this  time under development a  "Handbook  for Sampling and  Sample
       Preservation of Water and  Hastewater" by the U.S.  Environmental  Protection
       Agency.   When this  book  is  available it will serve  as  a good reference
       source on the topic of sampling.

       The Sampling section  of  the Criteria and Procedures  Document for Laboratory
       Certification spells  out the  mandatory requirements  that must be adhered to
       for the drinking water sampling.  It is attached  here  for student  reference
       as this  outline is  read.

       It has  been suggested that  one  of the certification  team members be from
       the staff of the  regional  water supply staff.   This  individual  would  be
       responsible to assess the  following:

       A.  Choice of sampling location
       B.  Proper sampling procedures
       C.  Sample identification
       D.  Prompt sample  transport to  the  laboratory
       E.  Sampling frequency
       F.  Bad-sample follow-up
       G.  Dissemination of  data  by  the laboratory  and use  of the data
           by  the water  supply  supervision  program

  II.   MONITORING REQUIREMENTS

       A.  Inorganic

           1.   Analysis for  all community water systems  utilizing surface water
               sources shall  be completed within  one  year  following June  1977.
               These analyses shall  be repeated at  yearly  intervals.

           2.   Analysis for  all community water systems  utilizing only  ground
               water  sources shall  be  completed within two  years  of June  1977.
               These analyses shall  be repeated at  three year intervals.

CH.MET.33. 3.30
                                                                          E2-1

-------
            3.   For  non-community water  systems, whether supplied by surface or ground
                water  sources,  analysis  for  nitrate  shall be completed within two years
                of June  1977.   These analyses shall  be repeated at intervals determined
                by the state or other  regulatory agency.
    Sample  collecting,  handling, and  preservation   — mandatory requirements
    Parameter

    Arsenic

    Barium

    Cadmium

    Chromium

    Lead

    Mercury


    Nitrate

    Selenium

    Silver

    Fluoride

    Chlorinated
     hydrocarbons


    Chlorophenoxys
  Preservative

Cone HN03 to pH<2

Cone HNCL to nH<2

Cone HN03 to pH<2

Cone HN03 to pH<2

Cone HN03 to pH<2

Cone HN03 to pH<2
Cone
           to pH<2
Cone HN03 to pH<2

Cone HN03 to pH<2

None

Refrigerate at 4°C
as soon as possible
after collection

Refrigerate at 4°C
as soon as possible
after collection
Container^- '
P or G
P or G
P or G
P or G
P or G
G
P
P or G
P or G
P or G
P or G
nth foil or
Maximum
holding time
6 months
6 months
6 months
6 months
6 months
38 days
14 days
14 days
6 months
6 months
1 month
14 days^
                                                                               (2)
                       Teflon-1ined cap
                       G with foil  or
                       Teflon-lined can
   (1)  P = Plastic, hard or soft, G = Glass, hard or soft
   (2)  In all cases, samples should be analyzed as soon after collection as
        possible.
   (3)  Well-stoppered and refrigerated extracts can be held up to 30 days.
   (4)  If a laboratory has no control over these factors, the laboratory director
        must reject any samples not meeting these criteria and so notify the authority
        requesting the analyses.
   (5)  If nitric acid cannot be  used because of shipping restrictions, sample may be
        preserved by icing.  Upon receipt  in the lab, the sample must be acidified with
        cone. HN07 to pH>2.  At time of analysis, sample container should be thoroughly
                 O
        rinsed with  1:1 nitric acid washings added  to the sample to  be  processed for
        subsequent analysis.
E2-2

-------
           4.  When the maximum contaminant level is surpassed, the frequency of
               resample shall be designated by the state and shall continue until
               the maximum contaminant level has not been exceeded in two successive
               samples or until  a monitoring schedule as a condition to a variance,
               exemption or enforcement action shall become effective,

       B.  Organic

           1.  For all community water systems utilizing surface water  sources,
               analysis shall be completed within one year of June 1977,   Samples
               analyzed shall be collected during the period of the year designated
               by the state as the period when contamination bv pesticides is most
               likely to occur.   These analysis shall be reoeated at intervals
               specified fay the State, but in no less frequency than at three
               year intervals,

           2.  For community water systems utilizing only ground water  sources,
               analysis shall be completed by those systems specified by the
               State.

           3.  If the results of an analysis indicate that the level  of any con-
               taminant exceeds  the maximum contaminant level, the supplier of
               water shall  report to the State within 7 days and initiate three
               additional  analyses within one month.

           4,  When the average  of four analyses exceeds the maximum contaminant
               level  the supplier of water shall  report to the State.   Horn'toring
               after public notification shall  be at a  frequency designated hy
               the State and shall  continue until  the maximum contaminant level
               has not been exceeded in two successive  samples or until  a moni-
               toring  schedule as a condition to a variance, exemption  or enforcement
               action  shall  become effective,

III.   SAMPLE  CONTAINERS

      A.   Types

          Generally two types of containers  are  acceptable;  these are glass  and
          plastic.   Plastic  is the  more convenient from a  shipping standnoint;
          however,  plastic ray not  be used  for  the organic  parameters.  The  glass
          containers  should  preferably be made  from a hard  borosilicate glass
          (Klmax  or Pyrex;  however,  other forms  may be  used).

          All  these various  materials have  certain advantages  and disadvantages.
          The hard  glass  is  inert to  most materials.  Conventional  polyethylene
          is  to be  used  when plastic  is acceptable because  of  reasonable  cost
          and less  adsorption.   Disposable  type  plastic containers,  such  as  the
          molded  polyethylene "Cubitainer,"  are  convenient  to  use.

          Usually,  a wide mouth  container is  praferred.  This  allows  easy sample
          removal  and  easier cleaning.

          Depending on the  State or  Regional  requirements,  three or  four  con-
          tainers will  be needed for  a  complete  analysis.   Usually,  a one qallon
          or  equivalent  size plastic  type container will suffice for  the  metals
          analyses.  A one  quart or  equivalent  plastic  container is  needed for


                                                                          E2-3

-------
    the nitrate sample.   A glass  container  holding  about  a  quart  with  a
    foil  or Teflon lined  screw cap is  required  for  analyses of  the
    organics,  and  finally, a  one  quart plastic  type container will  be
    needed for the fluoride parameter.

    Considering the section on frequency of monitoring,  the following
    are needed for each  public water supply in  one  years  time:

    1.   One 1  gallon plastic  container

    2.   Two 1  quart plastic container

    3.   One 1  quart glass container

    Additional containers would be needed for resampling  needs.   A  non-
    community  supply would probably need one 1  quart plastic container
    each year  unless resampling is to  be carried  out.

    In  addition to the containers themselves, some  type  of  shipping
    container  must be provided for each sample  container.   These  shipping
    containers can usually be purchased from the  supplier of the  actual
    sample container.

B.   Preparation and Shipment of Containers

    Individual responsibility to  provide, maintain  and clean sample con-
    tainers is dependent on how the State has elected to  carry  out  the
    certification program.  The laboratory or the authority could purchase
    in  large lots and make available sets of containers  to  each supply or
    the State  may elect to require the supply to  provide  their  own  con-
    tainers.  Generally speaking, the clastic containers  should not be
    reused for any trace analyses, that is, the metals.   The qlass
    containers should follow the  suggested  cleaning procedure,  including
    muffling  at 400°C for about  15 minutes.  Once  cleaned, these containers
    should be  stored and shipped  in such a  manner as to  prevent recontamination,

    Should the decision to reuse  plastic containers be made, they should
    be  cleaned carefully before reuse.  There are several cleaning  methods
    available.  Choosing the best method involves careful consideration
    of  the nature of the sample and of the constituent(s) to be determined.

    1.   Traces of dichromate cleaning solution will interfere with  metal
        analyses.   Use 1:1 nitric acid wash.

    2.   Traces of nitric acid may interfere with  the nitrate analysis.
        Use detergent with thorough rinses with tap and  distilled water.

    Shipping the containers to the sampling locations should take into
    consideration the numbers to  be shipped and eliminate any contamination
    chances.  The shipping containers to be used  in transporting  the sample
    itself to  the laboratory must be provided either as  a container for
    the empty  sample container or in bulk form.

-------
         One item that must be given consideration is the preservative.   Postal
         regulations will  not permit mailing of acids,  particularly nitric
         acid.   Consequently, these materials must be purchased locally  or
         shipped by truck  or other common carrier.  If  the materials are to be
         purchased locally, the purity must be riqidly  controlled to assure no
         contaminants are  present to effect results.

         Even when the preservative nitric acid has been  added  and diluted  by
         the sample, postal restrictions  may preclude the use of the mails.
         Therefore, a special footnote has been added to  the certification
         procedures allowing an alternate icing followed  by acidification upon
         receipt of the sample in the laboratory.

         When shipping the sample to the  laboratory for analysis,  sufficient
         time should be allowed to assure that the holding times are not sur-
         passed.   Alternate forms of transportation should be checked out before-
         hand to allow use if needed.   The sample  container must be protected
         from physical  damage in shipment and sufficient  coolant added to the
         ice chest or other form of insulated container to last through  the
         duration of shipment.   Caps should be checked  when the sample is taken
         to  assure that they will  not leak.   Upon  receipt in the laboratory,
         any deviation from the mandatory sampling requirements, i.e., preservative,
         holding  times, should  be noted and,  if necessary,  a resample ordered
         immediately.

IV.   SAMPLE  COLLECTING

     According  to the National  Interim Primary Drinking Hater Regulations,
     section 141.2(c),  the sampling location  is the "free-flowing  outlet of
     the  ultimate consumer."   Since this  represents a minimal effort, one
     sample  can be taken at any point  in  the  distribution system and  fulfill
     the  regulation.   Some States may  require more frequent samples  at randon
     locations,  or a  single composite  sample  taken at various locations.

     The  exception to  this sampling location  is the turbidity sample  which
     must be taken at  the  point of entry  of the water into  the  distribution
     system.

     When collecting  the sample,  the tap  should be run  to assure that the
     water collected  is from  the distribution system  and  not from  the private
     pipes.   The  sample container should  be flushed two or  three times before
     the  actual  sample  is  taken.   The  container should  not  be filled  completely
     to allow extra  volume for  effects of temperature during transit.  The
     preservative,  if  any,  should  be carefully added  to the container, the
     container  capped  and  the sample shaken.

     If the  sample is  to be cooled  during  shipment, the sample  container
     should  be  placed  in an insulated  container and sufficient  coolant added
     to last during  shipment.

     The  sample  should  be  labeled  to identify it during future  analyses.   The
     information  should include:

     A.   Date,  place  and time of  sampling;  name of person collecting  the
         sample .

-------
     B.   Identification of the sample as  to whether it is  a  routine  distribution
         system sample, check sample, raw or process water sample, or other
         special  purpose sample.

     C.   Analysis to be run on the sample as well  as any preservative added and
         what amount has been added.

     D.   Any other remarks that the sampler thinks is necessary.

     This information should be affixed to the sample container in such  a  way
     as  to assure that it will not become separated in later handlings.

     The Criteria and Procedures  Document for Laboratory Certification states
     that chain of custody procedures must be carried out  on all  samples taken
     for potential enforcement actions only.  The  exact procedure and directions
     on  this procedure should be  obtained from the appropriate certification
     authority.

 V.  FIELD MEASUREMENTS

     As  set down in the act, there are two types of analysis which may be
     carried out in other than certified  laboratories.  These are the analysis
     for residual chlorine and turbidity.  These measurements may be carried
     out in the field.  In addition,  should any other information about  the
     sample be required, such as  pH,  temperature,  etc., these should also  be
     carried out in the field.  It is not the scope of this  outline  to discuss
     the procedures involved in these analyses.   Procedures  on the residual
     chlorine and turbidity have  been included in  this manual for information
     purposes only and may be found under Tab E.

     State regulations may require additional procedures to  be carried out by
     the person taking the sample.  The Interim Primary Reflations  do not.

VI.  SUMMARY

     Proper sampling is the foundation of meaningful analytical results.  Con-
     sequently, a laboratory should know what constitutes  a  meaningful sample
     in  order to judge wiien a resample is necessary due to improper  sampling,
     preservation or handling techniques.

     The preservative to be added, the type of container and the holding times
     are spelled out in the Criteria  and  Procedures Document in a mandatory
     section.

     The laboratory Certification Officer must evaluate whether or not the
     laboratory is conducting a proper sample receipt procedure and, if  it has
     the responsibility, a proper sampling of the  v/ater supplies.

-------
                                 STATISTICS FOR CHEMISTS
 I  INTRODUCTION

 A  Statistics may be defined, for our purpose,
    as a collection of methods which have been
    developed for handling numerical data
    pertaining to samples or  portions of entire
    populations.

 B  The statistical methods with which we will
    concern ourselves deal with the presentation
    and analysis of numerical data from samples.
H   FREQUENCY

 A  Definitions

    1  Frequency - indicates how many times
      a particular score occurs in a collection
      of data
Frequency table - a tabular arrange-
ment of data,  ranked  in ascending or
descending order of magnitude,
together with the corresponding
frequencies

Frequency histogram - a set of
rectangles having bases on a horizontal
axia with centers at the given scores
and heights equal to the corresponding
frequencies (See Figure 1)

Frequency polygon  - a line graph of
frequencies plotted against scores
(can be obtained by connecting mid-
points  of tops  of rectangles in the
frequency histogram)   (See Figure 1)
                                             Figure 1

                            Frequency Histogram & Frequency Polygon
Frequency
— to w *» cj
11 1 1 1
X
X

X
X
^
X
^
/
/
/
S~

/v
t
•s
^N
S
"V
\
\
\
i,
\
\
V
1 I 1 I I
98 99 100 101 102
                                          Chloride
   ST. 25b. 11. 77
                                                                                       E3-1

-------
 Statistics For Chemists
B  Application

   Consider the application of the above
   definitions to the following set of data,
   obtained from twelve determinations for
   chloride in water.
               Results (ujE/1)
      100         101          99

      101         100         100

       99         102         100

       SB         101         102
                                               number of observations the median is
                                                 H
                                                 2
                                                          Xn +  X
                                                                ,  the average of the
                                               middle two scores,

                                            4  Mean - arithmetic average of all the
                                               values in the sample distribution, de-
                                               noted by X,  The formula for calcula-
                                               ting the sample mean is
                                              x -
                 Table 1

             Frequency Table
Chloride (]tg/l) _ ____________ Frequency
      98

      99

     100

     101

     102
                     1

                     2

                     4

                     3

                     2
III  MEASURES OF CENTRAL TENDENCY

 A Definitions

    1  Central tendency - the tendency of
       values to cluster about a particular
       value in the distribution

    2  Mode - that value which occurs most
       frequently

    3  Median - midpoint of an array of
       scores.  If there is an odd number of
       observations, n,  the median is
        X!L±JL   where  Xn + 1
                              2
 2
n+ 1
                                   represents
       the •'•'   J   value in the frequency

       distribution.   If ^ere is ^ even
                                                                   where there are n number
                                                                   of values.

                                                   B  Aids in calculation of the mean

                                                      Application of the following two statements
                                                      can reduce errors and amount of time
                                                      apent in calculating the mean of a
                                                      distribution.
                                                      1
Adding or subtracting a constant to or
from each score in a distribution is
equivalent to adding or subtracting the
same constant to or from the mean of
the distribution.  Thus the following
formula:
                                                       X  H X± C
                                                            where the X^'s are the
                                                       values in the distribution with mean X,
                                                       and the Xj ± C's are the_values in the
                                                       distribution with mean Xc.

                                                    2  Multiplying or  dividing each score  in
                                                       B distribution by a constant is equivalent
                                                       to multiplying or dividing the mean of
                                                       the distribution by the same constant.
                                                       Thus the following formulas:
                                                       (1) Xc = CX
                                                       or
                                                          xc =
                                                          where the X.'s are the
                                                       values in the distribution with mean X,

-------
                                                                Statistics for Chemists
   and the CX^s or the
                            's are the
          in the distribution with mean
C  Application

   Consider the application of the above
   definitions to the previously mentioned
   set of data,  obtained from twelve deter-
   minations for chloride in water, shown
   in Table 1.

   1  Mode = 100
2  Median =

   100 + 100
  2

= 100
                                               X  =
                                                c
                                      Denote the mean of the distribution in
                                      Table 1 by X  .  If we add 100 to each
                                      score in the distribution in Table 2, we
                                      obtain the scores in the distribution in
                                      Table 1; likewise if we add 100 to the
                                      mean, X,  of the distribution in Table 2,
                                      we obtain the mean, X , of the  distri-
                                      bution in Table 1.
                                                     Thus X  = X + 100
                                                           c
                                                          X  =-
                                                           c    n
                                                                      100
                                                    l(-2) + 2(-l) + 4(0) + 3(1) + 2(2)
                                                                 12
                                                                     + 100
                                               X   = .25 + 100 - 100.25
3  Mean =
           n
   98 + 2 (99) + 4 (1DO[ + 3 (101) 4- 2 (102)
                    12
=  100.25

4  Aid in Calculation

  Consulting Table 1 and observing that
  the values are in the neighborhood of
  100 we might subtract 100 from each
  score and obtain the following distribution:
               Table 2
            Frequency Table

Chloride C{*g/l)           Frequency
     -2                       1

     -1                       2

      0                       4

      1                       3
      2                       2
                                             IV  MEASURES OF DISPERSION

                                              A  Definitions

                                                 1  Dispersion - spread or variability of
                                                    observations in a distribution

                                                 2  Range - the difference between the
                                                    highest value and the lowest value

                                                    R  -  max - min

                                                 3  Average deviation - the sum of the
                                                    deviations of the values from their
                                                    mean, without regard to sign, divided
                                                    by the total number of data values (n)

                                                    The formula for calculating the average
                                                    deviation is:
                                                    d.
                                                                                   E3-3

-------
Statistics for Chemists
      Average deviation of the mean (D) -
      the average deviation of Individual
      data items  from the mean (d) divided
      by the square root of the number of
      data items  (n)

      The definition of the average deviation
      of the mean can be expressed by the
      formula:
               d
        D  =  , —  -
             •J  n
     Variance - the sum of the squares of
     the deviations of the values from their
     mean divided by the total number of
     data values (n) minus 1

     The definition of the- variance can be
     expressed by the following formula:
              n -  1
     Standard deviation - the square root
     of the variance

     The definition of the standard deviation
     can be expressed by the following
     formula;

     However, the formula commonly used
     because of its adaptability to the hand
     calculator is the following:
where there are
n number of values.
                              The definition of the standard
                              deviation of the mean can be
                              expressed by the formula:
                              S  =
  B

 / n
                              Relative standard deviation  - the
                              standard deviation (s) expressed as
                              a fraction of the mean,   s
                              The relative standard deviation is
                              often expressed as a  percent.  It is
                              then referred to as the  coefficient
                              of variation (V):

                              V  = ~  X  100 =  %
                                   X

                              The relative standard deviation is
                              particularly helpful when comparing
                              the precision of a  number of deter-
                              minations on a given  substance at
                              different levels of concentration.

                        B Aids in Calculation

                           Application of the following statements
                           can reduce errors and amount of time
                           spent in calculating the variance or
                           standard deviation of a distribution.
                                                    1
                              Adding or subtracting a. constant to or
                              from each score in a distribution
                              doesn't affect the variance or standard
                              deviation of the distribution.
                              Thus the following formulas:

                              (D

                              (2)
  2    2
8   = S
 C
             n - 1
     Standard deviation of the mean (S) - the
     standard deviation of individual data
     items (s) divided by the square root of
     the number of data  items (n)
S  « S
 C

where the X.'s are the values in
the distribution with variance s
and standard deviation s, and the
Xj +•'  C's are the values in the
distribution with  variance s  ^
and standard deviation s  . C
   E3-4

-------
                                                                   Statistics for Chemists
 2  Multiplying or dividing each score in
   a distribution by a constant is equivalent
   to multiplying or dividing the variance
   of that distribution by the square of the
   same constant.

   Thus the following formulae:

   (1)   82 = C2S2
         where the X;'s are the values in
                                        o
         the distribution with variance s ,
                              "V"
         and the CX.'s  or the  ..... _ i 's are
         the values in the distribution with
         variance s  ,

3  Multiplying or dividing each score in a
   distribution by a constant is equivalent
   to multiplying or dividing the standard
   deviation of that distribution by the
   same constant.

   Thus the following formulas:

   (1)    s  = Cs
   or    c
   (2)    .-£
        where the Xj's are the values in
        the distribution with standard
        deviation s, and the CX.'s or the
        "V" »
        r;|   's are the values in the
        C
        distribution with standard
        deviation s  ,
C  Application

   Consider the application of the above
   definitions to the previously mentioned
   set of data,  obtained from twelve
   determinations for chloride in water,
   shown in II B, Table 1.

   1  Range =» 102 -98 * 4
                                                                                    ~ X 1
4
n
1
2
4
3
2


Average aeviat;
Xi
98
99
100
101
102
X = 100.25
. _j:fXi~ x|
ion - Q -
|Xt- X|
2. 25
1.25
. 25
. 75
1. 75

11. 50
n
njXj^Xl
2. 25
2. 50
1.00
2.25
3. 50
11.50
- . 9fi
                                                                              12
                                                      3  Average deviation of the mean -
                                                         Using calculations from number 2,

                                                        D .  d * 0.96 = 0.96 = Q 2fl
                                                             ,—  .__.   3.46
                                                           •J n  s/12

n
1
2
4
3
2

.2 .

Xt
98
99
100
101
102



Xi-X
-2. 25
-1. 25
- .25
+ . 75
+1. 75

I- X)2
n-1
(Xi- X)2
5.06
1.56
. 06
. 56
3. 06

16.22

n(Xi- X)2
5.06
3. 12
. 24
1. 68
6. 12_
16.22
A1
                                                         n- i

                                                                                      E3-5

-------
 Statistics for Chemists

/


n Xj nXi X ,
1 98 98 9604
2 99 198 9801

4 100 400 10000
3 101 303 10201
2 102 204 10404
1203
/ 	 2
/120617--i222. /
_ - / 12 /120617
° V 11 4U
n - 1
nx!i
;:: .- AV-sf.ynr . ,„
7 n v/
40000 v n - 1
30603
20808 7 Standard deviation of the mean -
120617 s s

120801 Using calculations from number G,
1
s =
  ..21
                                                                        3.46
   6  Aid in calculation

      Recalling that adding or subtracting a
      constant to each score in the distri-
      bution doesn't affect the variance or
      the standard deviation of the distribu-
      tion we can simplify the computations
      by first subtracting 100 from each
      score in the distribution, thus obtain-
      ing the frequency distribution shown
      in Table 2.
   n
   1
   2

   4

   3
   2
xrc n(xrc)  oq-c)2  n(xrc)2
         EX-
            *
            n - 1
                                          8  Relative standard deviation expressed
                                            as a percent (coefficient of variation)

                                            V « -2-  X100
                                                 X
                                            Using calculations from number 6 for
                                            s = 1. 22 and from number 2 for
                                            X = 100.25,
                                                         Figure 2
                                                Normal Distribution Curve
                                                             Quantity Measured
E3-6

-------
                                                                  Statistics for Chemists
INTRODUCTION TO NOHMA L
DISTRIBUTION CURVE

Statistics deals with theoretical curves
which are smoother  than frequency
polygons, obtained from experiments in
real life.  However,  frequency distribu-
tions or frequency polygons of experimental
data often approximate a mathematical
function called the "normal" distribution
curve.  (See Figure 2)
                      As shown in Figure 3, the frequency polygon
                      for the 12 determinations for chloride in
                      water is a fairly good approximation of the
                      normal curve.  If,  however, in the chloride
                      determinations we  had obtained 103  instead
                      of 98 and 104 instead of 99 this distribution
                      would not have- been a good approximation of
                      the normal  curve,  as is shown in Figure 4.
                                       Figure 3
                  Comparison of Normal Curve and Frequency Polygon
    o
    a
    a*
   in
                •
                97
98      99      100      101

            Chloride wg/l
102
                                      Figure 4

                 Comparison of Normal Curve and Frequency Polygon
                                                                                   E3-7

-------
          for Chemists
B  If a frequency distribution is a good
   approximation of the normal curve, we
   can use some facte about the normal
   curve to give us information about the
   frequency distribution.
Figure 5 shows the normal distribution
in terms of the population mean u,  and
the  standard deviation of the population
o ,  and gives the percent of area under
the  curve between certain points.
                                         Figure 5

                                Normal Distribution Curve
                                           Figure 6

                                Frequency Distribution Polygon
    o
    c
    0)
    3
    O"
    Q)
    l-f
                                            /Tv



/
1


/ i'
^L f
' P !
-2s I -Is
98
/
/
i



i
99




>
1(
'


1
r
)0




I
101
J N
1 f.
1 j\
i it
+ ls * +2 a
102


k
V
\


                                      Chloride ng/1
  E3-8

-------
                                                                  Statistics for Chemists
 We may check the distribution of sample
 data to see if it is a "normal" distribution
 in the following manner.  Substitute the
 value of the sample mean (X) for the value
 of the midline and substitute the value of
 the sample standard deviation (s) for the
 limits of the value spans where we might
 expect certain percentages of the data
 items to occur.  Then we can check the
 number of data items which actually do
 occur within these value spans.

 Figure 6 demonstrates this application
 using the chloride data values from Table 1.
 The data values are marked on the hori-
 zontal line and the frequency of the
 occurrence of each value is marked on the
 vertical.  The midline of the distribution
 is marked at the value of the sample mean
 GT  = 100, See HI C 3).  The value of the
 sample standard deviation (s = 1.21,  See
 IV C 5) is used to mark value areas under
 the curve where different percentages of
 data values will probably occur.  Thus,
 for the area X + IB,  X - Is = 98. 79 and
 X + 1 s = 101.21.  Therefore, according
 to the normal distribution curve  shown in
 Figure 5, we might expect about 68% of the
 data items to have values between 99  and
 101.  (The values are rounded to whole
 numbers since the data values are thus
 recorded).

 Consulting Table 1,  we find that  75% or  9
 of the 12 data items have values in this
 range.  This percentage is  shown in
 Figure  6 by the frequency polygon for the
 data shown earlier in Figure 3.

 Likewise assuming a normal distribution,
 we would expect 95% of the observations
 to lie within +2 g 's from the population
 mean.   In fact, 100% of the observations
 were within + 2 s's from the sample mean.

 In both cases the observed percentages are
 reasonably close to the expected  percentages.
 Other tests exist for determining whether
 or not a frequency distribution might
 reasonably be assumed to approximate
the normal distribution.
    It would be good to become as familiar as
    possible with the normal distribution since
    an underlying normal distribution is
    assumed for many statistical tests of
    hypothesis.
 REFERENCES

 1  Bennett, C.A. and Franklin,  N. L.
      Statistical Analysis in Chemistry and
      the Chemical Industry.  John Wiley
      & Sons, Inc., New York.   1954,

 2  Crow, E. L., Davis, F.A., and Maxfield,
      M.W.   Statistics Manual.    Dover
      Publications, Inc., New York.   1960.

 3  Dixon, W.J. and Massey, F.J,
      Introduction to  Statistical Analysis.
      McGraw-Hill Book Co., Inc.,  New
      York.   1957.

 4  Ostle, B.   Statistics in Research.  The
      Iowa State University Press, Iowa.
      1963.
   Youdon, W.J.
      Chemists.
      New York.
 Statistical Methods for
John Wiley & Sons, Inc.,
 1951.
This outline was prepared by L. A, Lederer,
Statistician, formerly with Analytical
Reference Service, Training Program,
NCUIH, SEC.   Revised by Audrey  D.
Kroner, Chemist. National Training and
Operational Technology Center, MOTD, OWPO
USEPA, Cincinnati,  Ohio 45268.


Descriptors:  Graphic fVJethods, Quality Control,
Statistical Methods.  Statistics
                                                                                    E3-9

-------
                                  QUALITY CONTROL
    I.  INTRODUCTION
        The purpose of the Safe Drinking Water Act is to assure the public of an
        adequate supply of safe water.  To achieve this, maximum levels of certain
        contaminants were proposed along with the prescribed methodology for analyz-
        ing for these parameters.  When a laboratory performs these analyses, the
        laboratory should practice quality control to assure that the results being
        reported are true values and not in error.

        Data developed from these examinations must be reliable and beyond reproach.
        The data can be used for making judgments on technical operations in water
        treatment or in legal actions involving public health hazards.  For these
        reasons the U.S. Environmental Protection Agency in its Manual for the
        Interim Certification of Laboratories Involved in Analyzing, Public Water
        Suppl ie's has set down some required and some optional quality control
        procedures.

        The entire section contained in the "Manual for Certification" document is
        considered as the minimum acceptable program on quality control that a
        laboratory can carry out and still expect reliable results.  Most labora-
        tories will want to go beyond these minimum requirements and include more
        quality control.

        This outline will  cover the minimum quality control procedures, then go
        into the optional  portions and proceed further into some ideas not in the
        Manual  for Certification document.  The reasons for going further are to
        acquaint laboratory certification personnel with sufficient information
        to be able to evaluate whether the laboratory has complied with the mini-
        mum sections and allow the Certification Officer to recommend further
        procedures.  The topic of quality control from all  aspects in a laboratory
        is well covered -in the Handbook for An^y_tJ£aJ^Qu._a_1 vty Control in Water and
        Wastewater Laboratories produced by the EPA and available from the Office
        of Technology Transfer.  The Certification Officer should keep in mind that
        technical  assistance to the laboratory he is evaluating is of prime impor-
        tance because through this assistance he can upgrade the laboratory to
        produce better results.

        Assistance to state Certification Officers can be obtained from the regional
        certification authority, the Analytical  Quality Control Officer in the region,
        or from Environmental Monitoring and Support Laboratory in Cincinnati, Ohio.

        The Quality Control  section of the Criteria and Procedures document has been
        attached for the use of the trainee.

   II.  QUALITY CONTROL FOR CHEMICAL ANALYSIS

        A.  Minimum Requirements

            1.   All quality control data must be available for inspection.
CH.MET.con.8a.10.78

                                                                            E4-1

-------
             2,  Laboratory must analyze an unknown performance sample (when
                available) once per year for parameters measured.  Results must
                be within the control  limits established by EPA for each analysis
                for which the laboratory wishes to be certified.  If problems
                arise, appropriate technical assistance will be provided, and a
                follow-up performance  sample should be analyzed.

             3.  Minimum Daily Quality  Control

                a.  After a  standard reagent curve composed of a minimum of a reagent
                    blank and three standards has been prepared, subsequent standard
                    curves must be verified by use of at least a reagent blank and
                    one standard at or near the MCL.  Daily checks must be within +_ 10
                    percent  of original curve.

                b.  If 20 or more samples per day are analyzed, the working standard
                    curve must be verified by running an additional standard at or
                    near the MCL every 20 samples.  Checks must be within +_ 10 per-
                    cent of  original curve.

         B.   Optional  Requirements:  The following quality control requirements are
             optional.

             1.  Current service contract is in effect on all balances.

             2.  Class S weights are available to make periodic checks on balances.

             3.  Thermometer  certified  by the National Bureau of Standards (or one
                of equivalent accuracy) is available to check thermometers in ovens,
                etc.

             4.  Color standards or their equivalent are available to verify wave-
                length settings on spectrophotometers.

             5.  Chemicals dated upon receipt of shipment and replaced as needed or
                before shelf life has  been exceeded.

             6.  Criteria have been established for a laboratory analyzing supply
                samples other than its own:

                a.  Laboratory should  perform on a known reference sample (when
                    available) once per quarter for the parameters measured.  The
                    measured value should be within the control limit established
                    by EPA for each analysis for which the  laboratory wishes to be
                    certified.

                b.  At least one duplicate sample should be run every 10 samples, or
                    with each set of samples, to verify precision of the method.
                    Checks should be within the control limits established by EPA
                    for each analysis  for which the laboratory wished to be certified.

                c.  Standard deviation should be calculated and documented for all
                    measurements being conducted.

                d.  Quality  control charts or a tabulation of mean and  standard
                    deviation should be used to document validity of data on a
                    daily basis.
E4-2

-------
C.  The Minimal  Requirements

    1.  All quality control  data must be available for inspection.   This
        statement assures the availability of the data.   The person certifying
        the laboratory might wish to make use of these data to assure himself
        that the laboratory  is practicing quality control  and to what extent.
        After an amount of data have accumulated, it can serve as a record of
        a continuing type of quality control  rather than a sporadic, hit or
        miss type.   At any time, should there be questions on the reliability
        of any data, the quality control records will  be available  to show the
        reliability of the data produced during the time period in  question.

        The guidelines for data reporting recommend that the records of chemical
        analyses should be kept by the laboratory for  not  less than three years.
        It would seem prudent that all quality control  data be kept for a like
        period of time.

        Data required would  include a record  of the results of the  yearly per-
        formance sample, a standard curve for each method  the laboratory has  been
        certified for, the records showing a  check of  this curve daily or each
        time the analysis is carried out.   If the laboratory analyzes 20 or more
        samples  per day, records should include the value  of a standard run after
        every 20 samples. Again, this is for a minimal  program and it would  be
        well for laboratories to adopt at least the recommended procedures listed
        in the Manual for Certification document.

    2.  Laboratory  must perform on an unknown performance  sample once per year
        for parameters measured,

        In a minimal  program this yearly check sample  would be the  first
        external  indication  of a problem in a laboratory to the certifying
        authority.   The required daily quality control  data would not be
        sent to  the certifying authority.   If unacceptable answers  were
        obtained for one or  more parameters,  the laboratory would be asked to
        analyze  a follow-up  performance sample.   If continued problems existed,
        the certifying authority could offer  some form of  technical  assistance
        to rectify   the problem.  If the data is borderline or perhaps sporadic
        in nature,  the Certifying Officer might wish to  schedule his next visit
        at a time when the questionable analytical  method  is being  performed.

        The principal  state  laboratory, as well  as  local  laboratories, will be
        required to analyze  an unknown performance sample.   This sample will
        be provided by the regional  authority which will  certify that labora-
        tory in  each  state.   The U.S.  Environmental  Protection Agency also
        plans to make available to states  samples which  can be used as per-
        formance samples for local  laboratories  which  the  state has responsi-
        bility for  certifying.   The  performing laboratory  will  be given results
        of their analysis in terms of being within or  out  of the acceptable
        1imits.

        Results  must  be within the control  limits established by EPA for each
        analysis for  which the laboratory wished to be certified.

        The laboratory will  be informed if they  have or  have not complied with
        this requirement by  the authority  supplying the  performance sample.


                                                                    E4-3

-------
              3.   A standard curve must be prepared and kept for each  parameter the
                  laboratory analyzes  for.  This curve must be prepared with  a  minimum
                  of blank and three standards.   The references for the analytical methods
                  will  provide the laboratory with the range of the test.   Good procedure
                  would dictate choosing the three standards to cover  this  entire range.
                  A high, low and mid-range standard would be best to  run.   In  order to
                  assure good coverage the laboratory should be encouraged  to run more
                  than  the minimal requirements  as listed above.  A good recommended
                  procedure is to prepare the initial curve with a blank and  eight
                  standards covering the entire  range.

                  If more than one analyst will  run the same test, it  would be  wise  to
                  have  each analyst check their  procedural technique by checking the
                  standard curve.

                  After the initial curve has been established, it should be  verified
                  by the use of at least a reagent blank and one standard which has  a
                  concentration at or  near the MCL of the contaminant.   Again,  the Certi-
                  fication Officer should encourage more than the required  minimum
                  daily check.  The recommendation for good technique  recommends a
                  blank and two standards, one high and one low concentration.

                  These required daily checks of the standard reagent  curve should
                  be within +_ 10% of the original  concentration value.   For example,
                  if the MCL was 0.50, a standard  at this level analyzed as an  un-
                  known should fall between 0.55 and 0.45.  If not, the analyst
                  should check in the  following  order:

                  a.  Any variable instrument parameters

                  b.  Rerun check sample

                  c.  Prepare new standard

                  d.  Prepare all reagents fresh

                  e.  Check shelf life of chemicals

                  f.  Check instrument.

                  If the value persists at the new value through all this,  then the
                  analyst should prepare a new standard curve.

                  The laboratory Certification Officer should check to see  if the
                  daily records indicate that the curve has been checked (blank and
                  1 standard) and verified after each 20 samples with  a standard run.

          D.  Guidelines

              The following items are classed in the Criteria and Procedures  document
              as  recommended.  However, a certain amount of importance must be  attached
              to  each item.  The committee preparing the document felt strongly enough
              about these items to keep them in  the document.  Common  laboratory
              practice  would assure that these items be carried out.
E4-4

-------
1.  Current Service Contract on All Balances

    The analytical balance is of great importance in a laboratory.  As
    reagents are weighed on this piece of equipment, care must be taken
    to assure that it is in good working order.  The laboratory Certi-
    fication Officer should question the head chemist as to the existence
    of a service contract on the balances.  Should the laboratory Certi-
    fication Officer need additional information on proper care of a
    balance there is a section in the Handbook for Analytical  Quality
    Cgnt_rp1_m Water and Wastewater Laboratorjes,.

2.  Class S Weights Available to Make Periodic Checks on Balances.

    This could be included as part of the routine service contract or
    a set purchased and shared with the bacteriological  laboratory
    which will  also have need for them.  A very complete set of directions
    for checking the performance of a balance is contained in  Part 30 of
    ASTM Standards.

3.  NBS - Certified Thermometer Available to Check Thermometers in Ovens, etc.

    Again this  item could be a shared item between chemical  and bacteriological
    laboratories.  The Certification Officer could carry this  item with him
    and provide this service to the smaller type laboratories.   Since this
    item is only recommended,  the Certification Officer  can  only question
    if this thermometer is  avaialble and used.

4.  Color Standard or Their Equivalent Available to Verify Wavelength
    Settings on Spectrophotometers.

    The spectrophotometer like the balance is a very important  piece of
    laboratory  equipment.  The Certification Officers should make them-
    selves  thoroughly aware of the proper techniques for care,  use and
    calibration of a spectrophotometer.  Again the Handbook  for Analytical
    Quality Control  is a good  place to start.

    Spectrophotometers should  be checked for wavelength  alignment.   If a
    particular  colored solution is to be used at a closely specified wave-
    length, considerable loss  of sensitivity can be encountered if the
    wavelength  control  is misaligned.   In visual  instruments, an excellent
    reference point is the  maximum absorbance for a dilute solution  of
    potassium permanganate, which has  a dual  peak at 526 mu  and 546  my.
    On inexpensive graphing instruments,  which possess less  resolution
    than the prism instruments,  the permanganate peak appears at 525 to
    550 my  as a single flat-topped spike.

    Another point that should  be mentioned is the care and use  of spectre-
    photometric absorption  cells.   If possible,  the Certification Office
    should  observe the techniques  of the laboratory in the use  of the
    cells.   Good techniques here could indicate  good technique  in all
    the colorimetric procedures.
                                                               E4-5

-------
              5,  Chemical Dated Upon Receipt of Shipment and Replaced as Needed or
                  When Shelf Life is Exceeded.

                  It should not be necessary to store clean glassware or chemicals
                  on bench tops.  Floor length cabinets or above bench cabinets should
                  be available for storage.  Chemicals themselves should be of analytical
                  reagent grade to assure good quality.  Dating the chemical upon receipt
                  will give the chief chemist an Indication of the amounts to order and
                  if the chemical can still be relied on to have its initial quality.

              6.  Laboratories Analyzing Water Supply Samples Other Than Its Own
                  Should Carry Out Additional Quality Control.  This section covers
                 . additional optional items for the larger laboratories.

                  a.  Laboratory should perform on a known reference sample (when
                      available) once per quarter for the parameters measured.

                      Since the yearly known performance sample will not indicate to
                      the laboratory how well it is doing, other than pass or fail,
                      a known sample will show how the laboratory compares in precision
                      and accuracy to that given for the various methods.  Analysis of
                      the known sample will allow comparsion and show any trend of the
                      quality control of the laboratory.  These data should be available
                      to the Certification Officer for inspection.

                      This known quality control check sample should be available to
                      the laboratory from the principal state laboratory.  If not, a
                      synthetic sample prepared by the head chemist can be used.  This
                      control can be a large sample from a natural source known to
                      contain the constituents of concern or a synthetic sample prepared
                      in the laboratory from chemicals of the highest purity grade.  In
                      either case, if the control is to be kept, it should be stabilized
                      by addition of a suitable preservative.  See the section on
                      sampling for the choice of preservative.

                  b.  The measured value should be within the control limits established
                      by EPA for each analysis for which the laboratory wishes to be
                      certified.

                      Precision data can be found in one or the other standard
                      references.  That is

                      1)  Standard Methodsforthe Examinationof Water and Wastewater,
                          13th Edition (1971).""~

                      2)  Manual of Methods for the Chemical Analysis of Waterand Wastes,
                          1974 Edition.

                      These data have been accumulated in Table I.  If this data does
                      not fulfill the need of the Certification Officer, he may write to
                      the U. S. Environmental Protection Agency, EMSL, Cincinnati, Ohio
                      45268 and request additional information on accuracy and precision.
E4-6

-------
 c.   At  least  one  duplicate  sample should  be  run every  10 samples,
     or  with each  set  of  samples, to verify the precision of the
     method.   Checks should  be within  the  control  limits established
     by  EPA for each analysis for which  the laboratory  wishes to be
     certified.

     In  order  to document that reproducible results are being obtained
     (i.e. precision of the  method), it  is necessary to run duplicate
     samples.  Although the  frequency  of such replicate analysis is,
     by  nature dependent on  such factors as the original precision
     of  the method, the reliability of the instrumentation involved
     and the experience of the analyst, good  laboratory technique is
     to  run duplicate  analysis at least ten percent of  the time.
     The resulting data should be within the control limits established
     by  EPA.   If the data do not agree, the system is not under con-
     trol, and results are subject to question.

d.   Standard deviation should be calculated and documented for all
     measurements being conducted.

     This calculation will provide the upper and lower control  limits
     for the test.   Analysts can then determine whether or not the
     data produced  is acceptable.  This data can be calculated  on
     seven replicate determinations  for initial comparsion.   However,
     as additional  determinations are performed, they should  be added
     to existing data and the precision data recalculated.   Twenty
    or  more runs  tend to present better statistical  data.

    Standard deviation calculations  should be determined for each
    analyst to carry out the analysis.  However,  the data should
     not be collected until  the  analyst is  familiar with the  pro-
    cedure.   The concentration  used  to calculate  the standard
    deviation  should be at the  level  expected in  the sample  for those
     laboratories  doing only their own  water.   For  laboratories
    doing determinations other  than  their  own supply it would  be
    best to have the standard deviation calculated at several
    concentrations.  However, for a minimal effort, the concentration
    should be  chosen at or close to  the maximum contaminant  level
    for the parameter.

    In order to assure this  data is  collected, the standard  run after
    each 20 samples  could be at the  concentration  used to determine
    the standard deviation.   This would produce a  constant flow of
    this data  for  inclusion  in  future  updates of  the standard  devia-
    tion calculation.

e.  Quality  control  charts  or a  tabulation of mean and standard
    deviation  should be used to  document the  validity of data  on
    a daily basis.

    If the upper and lower  control  limits  of  t 2  standard  deviations
    are calculated,  the analyst will  have  some idea as to  the
    acceptability  of each determination as the results are obtained.
    When outliers  are  found  the  analyst can reschedule these  for
    analysis  to assure themselves of the result before action  is
    taken to call  for  a resample of  the supply.
                                                          F4-7

-------
                      Production of quality control charts and subsequent graphing
                      of  the charts of data obtained in the laboratory will give
                      pictorial representation of  the control of the method.  Ten-
                      dencies  toward one or the other control limit will indicate
                      loss of  control of the method.

                      How to produce quality control charts and a discussion of these
                      statistical  tools is covered  in the Basic Statistics outline
                      and in the Handbook for Analytical Quality Contrpj inWater and
                      Wastewater Laboratories.
E4-8

-------
                                TABLE  I
Maximum
Contaminant
Level ug/1
50


1000


10

50


50

2


10


50


10,000



Varies with
Temperature





Parameters
Arsenic
{Gaseous Hy-
dride)
Barium
(Standard
cond, )
Cadmium
(Extracted)
Chromium
(Standard
cond.)
Lead
(Extracted)
Mercury
(Flameless
AA Inst.)
Selenium
(Gaseous Hy-
dride)
Silver
(Standard
cond.)
Nitrate
(Brucine)
Cadmium Red

Fluoride
(SPADNS
with
Distil)
Electrode


Relative
Standard Standard
Cone. Deviation Deviation
W9/1 s ug/1 RSDf Reference
20.0 + 1.1
10.0

40.0 43
1000
500
10
50
74 29
50
50
50
50
3.4 1.49
0.4

10 1.1
10

50


5000

1040 10
5000
570
570 1 30


750
900
750 36

6.0


8.9
10.0
72.8
43.8

26.4
26.4
23.5
23.5

21.2


11.0

17.5


15.4


9.2
17.2



4.8
2.9

E.P
Std

E.P
Std
Std
Std
Std
E.P
Std
Std
Std
Std
E.P
Std

E.P
Std

Std


Std

E.P
Std
Std
E.P


Std
Std
E.P
.A. Methods - 95 9
. Methods (14th) 146^

.A. Methods - 98
. Methods (13th) 215
. Methods (14th) 146
. Methods (13th) 213
. Methods (14th) 148
.A. Methods - 106
. Methods (13th) 215
. Methods (14th) 146
. Methods (13th)
. Methods (14th)
.A. Methods - 125
. Methods (14th)

.A. Methods
. Methods (14th)

. Methods (13th)


. Methods (13th)

.A, Methods - 206
. Methods (13th) 464
. Methods (13th)
.A. Methods - 59


. Methods (13th)
. Methods (13th)
.A. Methods -67
Where more than one concentration and Standard Deviation is given in the same
reference the closest to the maximum contaminant level  has been given.

Although not an official reference, data included here.
                                                                   E4-9

-------
   III.   SUMMARY
         the  quality  control  items  in  the  Criteria  and  Procedures  document  identify a
         minimal  effort for all  types  of laboratories.   Since  quality  control  is for
         the  benefit   of the laboratory in assuring valid  data,  it would seem  wise for
         all  laboratories  to practice  a good  deal more  quality control  than set down
         in  the  Criteria Procedures  document.

         This section has  discussed  the quality  control  steps  to be  taken to assure
         proper  analytical  performance in  the laboratory.  However,  a  complete picture
         of  quality control  would  include  adherence to  proper  sampling techniques, in-
         cluding collection,  preservation  and handling;  use of acceptable methods, and
         proper  reporting  of data  to be considered.   It  must be  recognized  (and practiced),
         however,  that quality control begins with  collection  and  does  not  end until
         resulting data are reported.
E4-10

-------
                              LABORATORY SAFETY PRACTICES
 I  INTRODUCTION

 A  Safe Use, Handling and Storage of Chemicals

    I Chemicals in any form can be safely
      stored, handled, and used if their
      hazardous physical and chemical
      properties are fully understood and the
      necessary precautions, including the
      use of proper safeguards and personal
      protective equipment are observed.

    2  The management of every unit within a
      manufacturing establishment must give
      wholehearted support to a well integrated
      safety policy.

 B  General Rules for  Laboratory Safety

    1  Supervisory personnel should think
      "safety, "  Their attitude toward fire
      and safety standard practices is reflected
      in the behavior of their entire staff.

    2  A safety program is only as strong as
      the worker's will to  do the correct
      things at the right time.

    3  The fundamental weakness of most
      safety programs lies in too much lip
      service to safety rules and not  enough
      action in putting them into practice.

    4  Safety practices should be practical and
      enforceable.

    5  Accident prevention  is based on certain
      common standards of education, training
      of personnel and provision of safeguards
      against accidents.
II   LABORATORY DESIGN AND EQUIPMENT

 A  Type of Construction

    1  Fire-resistant or noncombustible

    2  Multiple story buildings should have
      adequate means of exit.
   3 Stairways enclosed with brick or
     concrete walls

   4 Laboratories should have adequate exit
     doors to permit quick,  safe escape in
     an emergency and to protect the
     occupants from fires or accidents in
     adjoining rooms.  Each room should be
     checked to make sure there is no
     chance  of a person being trapped by
     fire, explosions, or release of  dangerous
     gases.

   5 Laboratory rooms in which most of the
     work is carried out with flammable
     liquids  or gases should be provided
     with explosion-venting windows.

B  Arrangement of Furniture and Equipment

   1 Furniture should be arranged for
     maximum utilization of available space
     and should provide working conditions
     that are efficient and safe.

   2 Aisles between benches should be at
     least 4  feet wide to provide adequate
     room for passage of personnel and
     equipment.

   3 Desks should be isolated from benches
     or adequately protected,

   4 Every laboratory should have an eye-
     wash station and a safety shower.

C  Hoods and Ventilation

   1 Adequate hood facilities should  be
     installed where work with highly toxic
     or highly flammable materials are used.

   2 Hoods should be ventilated separately
     and the  exhaust should be terminated
     at a safe distance from the building,

   3 Make-up air should be supplied  to
     rooms or to hoods to replace the
     quantity of air exhausted through the
     hoods.
 PC.SA.lab. 1. 11.77
                                                                                        E5-1

-------
Laboratory Safety Practices
   4 Hood ventilation systems are best
     designed to have an air flow of not less
     than 60 linear feet per minute across
     .the face of the hood, with all doors open
     and 150, if toxic materials are involved.

   5 Exhaust fans should be  spark-proof if
     exhausting flammable vapors and
     corrosive resistant if handling corrosive
     fumes.

   6 Controls for  all services should be
     located at the front of the hood and
     should be operable when the hood door
     is closed,

   7 All laboratory rooms should have the
     air changed continuously at a rate
     depending on the materials being
     handled,

D  Electrical Services

   1 Electrical outlets  should be placed
     outside of hoods to afford easy access
     and thus protect them from spills and
     corrosion by gases.

   2 Noninterchangeable plugs should be
     provided for  multiple electrical services.

   3 Adequate outlets should be provided and
     should be of the three-pole type to
     provide for adequate grounding.

E  Storage

   1 Laboratories should provide for adequate
     storage space for mechanical equipment
     and glassware which will be used
     regularly,

   2 Flammable solvents should not be stored
     in glass bottles over one liter in size.
     Large quantities should be stored in
     metal safety  cans.  Quantities  requiring
     containers larger than one gallon should
     be stored outside the laboratory,

   3 Explosion proof refrigerators should be
     used for the  storage of  highly volatile
     and flammable  solvents.
   4 Cylinders of compressed or liquified
     gases should not be stored in the
     laboratory,

F  Housekeeping

   1 Housekeeping plays an important role
     in reducing the frequency of laboratory
     accidents.  Rooma should be kept in a
     neat orderly condition.   Floors,  shelves,
     and tables should be kept free from
     dirt  and from all apparatus and chemi-
     cals not in use.

   2 A  cluttered laboratory is a dangerous
     place to work. Maintenance of a clean
     and orderly work space  is indicative of
     interest,  personal pride, and safety -
     mindedness.

   3 Passageways should be kept clear to all
     building exits and stairways.

   4 Metal containers should be provided  for
     the disposal of broken glassware and
     should be properly labeled.

   5 Separate approved waste disposal cans,
     should be provided for the disposal of
     waste chemicals.

   6 Flammable liquids not miscible with
     water and corrosive materials, or
     compounds  which are likely to give off
     toxic vapors should never  be poured
     into the sink,

G  Fire Protection

   1 laboratory personnel should be
     adequately trained regarding pertinent
     fire hazards associated with their work.

   2 Personnel should  know rules of fire
     prevention and methods of combating
     fires.

   3 Fire extinguishers (CO  type)  should
     be provided at convenient locations and
     personnel should be instructed in their
     use.
   4 Automatic sprinkler systems are
     effective  for the control of fires in
     chemical laboratories.
E5-2

-------
                                                                 Laboratory Safc;ty Practices
 H  Alarms
       An approved fire alarm system should
       be provided.

       Wherever a hazard of accidental release
       of toxic gases exists, a gas alarm
       system to warn occupants to evacuate
       the building should be provided.

       Gas masks of oxygen or compressed air
       type should be located near exits and
       selected personnel trained to use them.
Ill  HANDLING GLASSWARE

 A  Receiving, Inspection and Storage

    1  Packages containing glassware should
       be opened and inspected for cracked or
       nicked pieces,  pieces with flaws that
       may become cracked in use, and badly
       shaped pieces.

    2  Glassware should be  stored on well-
       lighted stockroom shelves designed and
       having a coping of sufficient height
       around the edges to prevent the pieces
       from falling off.

 B  Laboratory Practice

    1  Select glassware that is  designed for the
       type of work planned,

    2  To cut glass tubing or a  rod, make a
       straight clean cut with a cutter or file
       at the point where the piece is  to be
       severed. Place a towel  over the piece
       to protect the hands and  fingers, then
       break away from the body.

    3  Large size tubing is cut  by  means of a
       heated nichrome wire looped around the
       piece at the point of severance.

    4  When it is necessary to insert  a piece
       of glass tubing or a rod through a
       perforated rubber or cork stopper,
       select the correct bore so that the
       insertion can be made without excessive
       strain.
    5  Use electric mantels for heating
       distillation apparatus,  etc.

    6  To remove glass splinters,  use a
       whisk broom and a dustpan.  Very
       small pieces can be picked up with a
       large piece of wet cotton.
IV  GASES AND FLAMMABLE SOLVENTS

 A  Gas Cylinders

    1  Large cylinders must be securely
       fastened so that they cannot be dis-
       lodged or tipped in any direction.

    2  Connections,  gauges,  regulators or
       fittings used with other cylinders must
       not be interchanged with oxygen
       cylinder fittings because of the possi-
       bility of fire or explosion from a
       reaction between oxygen and residual
       oil in the fitting.

    3  Return empty cylinders promptly with
       protective caps replaced.

 B  Flammable Solvents

    1  Store in  designated areas well
       ventilated.

    2  Flash point of a liquid is the temperature
       at which it gives off vapor sufficient  to
       form an  ignitible mixture with the air
       near the surface of the liquid or  within
       the vessel used.

    3  Ignition temperature of a substance is
       the minimum temperature required to
       initiate or cause self-sustained com-
       bustion independently of the heating or
       heated element.

    4  Explosive or flammable limits.   For
       most flammable liquids, gases and
       solids there is a minimum concentration
       of vapor in air or oxygen below which
       propagation of flame does not occur on
       contact with a source of ignition.
       There is also a  maximum proportion of
       vapor or gas in  air above which
                                                                                        E5-3

-------
 Laboratory Safety Practices
      propagation of flame does not occur.
      These limit mixtures of vapor or gas
      with air, which if ignited will just
      propagate flame,  are known as the
      "lower and higher explosive or flammable
      limits."

      Explosive Range.  The difference
      between the lower and higher explosive
      or flammable limits,  expressed in
      'terms of percentage of vapor or gas in
      air by volume is known as the "explosive
      range,"

      Vapor Density is the relative density
      of the vapor as compared with air.

      Underwriter's Laboratories Classification
      is  a standard classification for grading
      the relative hazard of the various
      flammable liquids.  This classification
      is  based on the following scale:

      Ether Class	    100
      Gasoline Class	  90 -  100
      Alcohol (ethyl) Class	  60-   70

      Kerosene Class	  30 -   40

      Paraffin Oil Class . .	  10 -   20
    8  Extinguishing agents


V   CHEMICAL HAZARDS

A  Acids and Alkalies

    1  Some of the most hazardous chemicals
      are the "strong" or "mineral" acids
      such as hydrochloric, hydrofluoric,
      sulfuric and nitric.

    2  Organic acids are less hazardous
      because of their comparatively low
      ionization potentials.  However, such
      acids as phenol (carbolic acid),
      hydrocyanic and oxalic are extremely
      hazardous because of their toxic
      properties.

    3  Classification of acids
B  Oxidizing Materials

   1 Such oxidizing agents as chlorates,
     peroxides, perchlorates and perchloric
     acid,  in contact with organic matter
     can cause explosions and fire.

   2 They are exothermic and decompose
     rapidly, liberating oxygen which reacts
     with organic compounds.

   3 Typical hazardous oxidizing agents are:

     Chlorine Dioxide
     Sodium Chlorate
     Potassium Chromate
     Chromium Trioxide
     Perchloric Acid

C  Explosive Power

   1 Many chemicals are explosive  or form
     compounds that  are explosive and
     should be treated accordingly.

   2 A few of the  more common examples
     of this class of hazardous  materials are;

     Acetylides
     Silver Fulminate
     Peroxides
     Peracetic Acid
     Nitroglycerine
     Picric Acid
     Chlorine and Ethylene
     Sodium Metal
     Calcium  Carbide

D  Toxicity

   1 laboratory chemicals improperly
     stored or handled can  cause injury to
     personnel by virtue of their toxicity.

   2 Types of exposure.  There are four
     types of exposure to chemicals:
     a  Contact with the skin and eyes
     b  Inhalation
     c  Swallowing

     d  Injection
E5-4

-------
                                                                 Laboratory Safety Practices
VI  PRECAUTIONARY MEASURES

 A  Clothing and Personal Protective Equipment

    1 'Chemical laboratories should have
       special protective clothing and equipment
       readily available for emergency use and
       for secondary protection of personnel
       working with hazardous materials.

    2  Equipment should be provided for adequate:

       a  Eye protection

       b  Body protection

       c  Respiratory protection
       d  Foot protection

       e  Hand protection

 B  Bodily Injury

    1  Burns,  eye  injuries, and poisoning are
       the injuries with which laboratory
       people must be most concerned.
   2 First emphasis in the laboratory
     should be on preventing accidents.
     This means observing all recognized
     safe practices using necessary personal
     protective equipment and exercising
     proper control over poisonous sub-
     stances at the source of  exposure.

   3 So that a physician can be summoned
     promptly, every laboratory should have
     posted the names, telephone numbers,
     and addresses of doctors to be called
     in an emergency  requiring medical care.


REFERENCES

Guide for Safety in the Chemical  Laboratory,
   the General Safety  Committee  of the
   Manufacturing Chemists Association,  Inc.,
   Van Nostrand.  New York (1954).
This outline was prepared by Paul F. Hallbach,
Chemist,  National Training and Operational
Technology Center,  MOTD, OWPO,  USKPA,
Cincinnati, Ohio 45268

Descriptors:  Safety, Laboratory, Practices
Safety,  Laboratory Design Chemical Storage,
Gas Cylinders, Flammable  Solvents
                                                                                      E5-5

-------
                            A  PROTOTYPE  FOR DEVELOPMENT OF
                            ROUTINE  OPERATIONAL PROCEDURES

                                      for

                            DETERMINATION OF SILVER  (Ag+)
                                   as applled in

                     WATER AND WASTEWATER  TREATMENT FACILITIES

                                   and in  the

                        MONITORING OF EFFLUENT WASTEWATERS
             National Training and Operational Technology Center
                      Office of Water Program Operations
                     U.S.  Environmental Protection Agency
CH.HE.ag.lab.WMPJ.il.77
                                                                        E6-1

-------
WATER MONITORING PROCEDURE:  Determination of Ag+

1. Analysis Objective:

   To determine the silver concentration of a water sample.

2. Brief Description of Analysis:

   The sample is digested with concentrated nitric acid and evaporated to
   dryness.  The residue is treated with hydrochloric acid, silicates and
   other insoluble material are removed by filtration and the sample is
   analyzed for the total metal of interest by atomic absorption spectro-
   photometry.

3. Applicability of this Procedure:

   The method works for both potable and wastewater.

   a. Range of Concentration - The method is  recommended for use in the
      range of 0.1  to 4.0 mg/1.  The detection limit  is 0,01 mg/1.
   b. Pretreatment of Sample - Digestion in acid pH to assure solubilization -
      See Section A.
   c. Treatment of Interferences in the Sample - None listed for these
      conditions.
Source of Procedure:   Methods for Chemical  Analysis  of Water and Wastes,  1974,
Environmental  Protection Agency,  Environmental  Monitoring  and Support Laboratory,
Cincinnati, Ohio, p.  146

                                                                      E6-3

-------
    WATER MONITORING PROCEDURE:   Determination of Ag+






    Operating  Procedures:



    A.  Sample  Digestion



    B.  Reagent Preparation



    C.  Instrument  Calibration



    D.  Instrumental Analysis



    E.  Calculations
E6-4

-------
WATER MONITORING PROCEDURE:   Determination of Ag+


General Description of Equipment and Supplies Used in the Process

A. Capital Equipment

   1. Balance, analytical  -  sensitivity 0.1  milligram
   2. Atomic absorption spectrophotometer and recorder
   3. pH meter
   4. Hot plate, 110 V

B. Reusable Supplies

   1. Flasks, volumetric,  100 ml, 1000 ml
   2. Pipets, volumetric,  50 ml, 3 ml, 1  ml
   3. Reagent bottles, glass with glass stopper
   4. Anion and cation exchange  resin cartridges
   5. Beakers, 150 ml
   6. pH paper
   7. Watch glass
   8. Funnel, 80 mm diameter
   9. Ring stand and 3 inch  ring
  10. Graduated cylinders  100, 50, 10 ml

C. Consumable Supplies

   1. Reagents

      Silver Nitrate (analytical reagent  grade)
                                                                      E6-5

-------
 WATER MONITORING PROCEDURE:   Determination of Ag+
                                                                                                        E6-6
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Sample Digestion
   for Total Silver
1. Transfer 100 ml of sample
   into a clean 150 ml beaker

2, Check the pH using
   pH paper
                           3. Add 5.0 ml  of 1:1
                              hydrochloric acid (HC1)

                           4. Place the beaker on a hot
                              plate

                           5. Heat at 95° C for 15 min

                           6. Remove the beaker from the
                              hot plate.   Allow it to
                              cool to room temperature

                           7. Wash down the beaker walls
                              with distilled water

                           8. Filter the sample through
                              Whatman $42 filter paper
                              into a clean 100 ml
                              volumetric  cylinder

                           9. Dilute the volume to 100
                              ml  with distilled  water
la. Use a 100 ml  graduated cylinder
2a. The pH should be 2.0,   If the sample was  not
    acidified upon collection, add 1:1  nitric acid
    dropwise until the pH is adjusted to 2.0

3a. Use a 5 ml  pipet. Use a rubber bulb on the pipet
                                4a.  Adjust the hot plate for medium heat
                                5a.  Make certain that the sample  does  not boil
                                7a.  Use a plastic wash  bottle
B. Reagent Preparation
   1. Dei om* zed
      Distilled Water
1.  Prepare by passing
   distilled water through a
   mixed bed of cation and
   anion exchange resions
la.  Use deionized distilled water for  the  preparation
    of all  reagents,  calibration  standards and  as
    dilution water.                                  i

-------
 WATER MONITORING PROCEDURE;    Determination  of  Ag+
 OPERATING PROCEDURES
        STEP SEQUENT
     INFORMATION/OFF RAT ING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B. Reagent Preparation
   (Continued)
   2. Nitric Acid
      Concentrated
      (HN03)

   3. Hydrochloric Acid
      (HC1) 1:1
   4. Silver Stock
      Standard Solution
   5. Fuel and Oxidant
1.  Commercially available
   reagent grade
1.  Prepare a 1:1  solution of
   reagent grade  hydrochloric
   acid by adding 25 ml  of
   commercially available
   reagent grade  hydrochloric
   acid to 25 ml  of deionized
   water

1,  Carefully weigh 1.575
   grams of silver nitrate
   (analytical reagent grade)
   on an analytical  balance

2.  Transfer into  a 1000 ml
   volumetric flask
3. Dissolve in distilled
   water

4. Add 10 ml  concentrated
   nitric acid

5. Dilute to  mark with
   distilled  water

1. Commercial  grade acetylene
   is generally acceptable
la.  Use a 50 ml  graduate
la. Use a plastic weighing boat and an analytical
    balance
                                                          2a.  Use  a  powder  funnel
                                                          2b.  Use  a  plastic wash  bottle  to  rinse  the weighing
                                                              boat and  funnel  into  the  flask
                                                          4a.  Use  a  graduate  cylinder
                                                          4b.  Use  caution with  the  acid
                                                          5a.  One  ml  equals  1  mg  Ag+(1000  mg/liter)
                                                                                                          E6-7

-------
 MATER MONITORING PROCEDURE:     Determination of Ag+
                                                                                                         E6-8
  OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B. Reagent Preparation
   (Continued)
   Air may be supplied from a
   compressed air line, a
   laboratory compressor, or
   from a cylinder of
   commercial air.
2a. Caution:   Air supply must  be  free  from oil
              or other contaminants
 C.  Instrument  Calibration
1 . Turn on air supply

2. Turn on acetylene supply

3. Turn on instrument and
   ignite flame

4. Turn on power to hollow
   cathode lamp

5. Select wave length for
   appropriate metal

6. Prepare a series of stand-
   ard solutions for silver
   as follows:

   Silver
   Transfer 1.0 ml  of stock
   silver solution  into a
   100 ml volumetric  flask
   and dilute to the mark
   with deionized distilled
   water and shake  well
                                                           3a. See instruction manual  for your particular
                                                               instrument

                                                           4a. Select lamp for proper metal  analysis
                                                           5a. Silver (328.1  nm)

-------
WATER MONITORING PROCEDURE:   Determination  of Ag+
 OPERATING  PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C.  Instrument Calibration
   (Continued)
6. Continued

   Transfer 0.0, 0,2, 0.4,
   0.6, 0.8, and 1.0 ml  of  thi
   diluted standard solution
   into each of six 100  ml
   volumetric flasks re-
   spectively.  Dilute to the
   mark with deionized dis-
   tilled water and shake we!
   The concentration of  these
   solutions will be 0.00,
   0.02, 0.04, 0.06, 0.08,  and
   0.10 mg/1 respectively

7. Ignite flame and aspirate
   standard solutions into
   the flame

8. Prepare a calibration curv
   by plotting the concentra-
   tion of the respective
   metals against the respons
   for each concentration
                                                               Use  a  1  ml  micro  pipet  graduated  in  0.1  ml
                                                           8a.  Record  the  response  on  a  recorder  or  use  the
                                                               readout provided  on  the instrument
D. Instrumental  Analysis
1. Aspirate the unknown solu-
   tion into the instrument
   immediately following the
   aspiration of the standard

2, Record the response
la. Flame characteristics and instrumental  settings
    should be the same for standards and unknowns.
                                                                                                           E6-9

-------
 WATER MONITORING PROCEDURE:  Determination of Ag+
                                                                            E6-10
 OPERATING PROCEDURES
     STCP SEQUENCE:
[JJFORKATION/OPCRATiriG GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E. Calculations
Determine the concen-
tration of the metal in the
sample by substituting the
observed instrumental re-
sponse on the appropriate
calibration curve.

-------
                       A PROTOTYPE FOR DEVELOPMENT OF
                       ROUTINE OPERATIONAL PROCEDURES

                                  for the

                DETERMINATION OF TOTAL CADMIUM, CHROMIUM AND
                        LEAD BY ATOMIC ABSORPTION
                                as applied in

                           WATER TREATMENT PLANTS
                                 and in the
                     MONITORING OF DISTRIBUTION SYSTEMS
             National Training and Operational Technology Center
                      Office of Water Program Operations
                     U.S. Environmental  Protection Agency


CH.ME.lab.WMP.la.A.SO
                                                                       E7-1

-------
WATER MONITORING PROCEDURE:   Determination of Total  Cadmium,  Chromium and
                             Lead by Atomic Absorption

1. Analysis Objectives:

   The user of the attached  procedure will determine the cadmium, chromium
   and lead content of a drinking water sample,  including sample preparation
   and atomic absorption.

2. Brief Description of Analysis

   If suspended or settleable matter is present, the sample is treated with
   acid and heat to assure complete solubilization of metals.   To determine
   total chromium, trivalent chromium is oxidized to the hexavalent form.
   The metals are chelated and extracted with pyrrolidine dithiocarbamic
   acid (PDCA) in chloroform.  An acidified water solution of the metals
   is aspirated into an atomic absorption spectrophotometer.

3. Applicability of the Procedure:

   a. Range of concentration:

      This procedure should  be carried out if the concentrations in the
      sample are below:

      0.-.P2CL JffiL/1 jfor cadmium 	
      0.~Q"5Q mg/1  for_chrom1um_
      0.200 mg/V for Tead

   b. Pretreatment of samples:

      This is covered in the procedure.  Generally it is to lower the pH of
      the sample to below pH 2 with nitric add  for preservation.  A solubiliza-
      tion procedure for "total" metals if particulates are in the sample
      is also covered.

   c. Treatment of interferences in samples:

      The method contains steps to remove interferences, i.e., chelation
      and extraction.  A section about interferences to atomic absorption
      spectrophotometry (chemical, dissolved solids, ionization and
      spectral) can be found in the Source of Procedure.*
*Source of Procedure:  Methods for Chemical  Analysis of Water and Wastes, 1974
 and 1979, U.S. Environmental  Protection Agency, Environmental  Monitoring and
 Support Laboratory, Cincinnati, Ohio 45268, Metals (Atomic Absorption Methods)
                                                                          E7-3

-------
    WATER MONITORING PROCEDURE:   Determination of Total  Cadmium,  Chromium and
                                 Lead by Atomic Absorption
    A.   Glassware Preparation
    B.   Sample Preservation and Handling
    C.   Reagent Preparation
    D.   Instrument Set-up
    E.   Solubilization for "Total" Metals (If necessary)
    F.   Preparation of Standard Dilutions
    G.   Oxidation for Total Chromium
    H.   Extraction of Metals
    I.   Instrument Calibration
    J.   Calculations
    K.   Instrument Shut-Down
    L.   Maintenance
E7-4

-------
WATER MONITORING PROCEDURE:
              Determination of Total  Cadmium,  Chromium and
              Lead by Atomic Absorption
Equipment and Supply Requirements

A. Capital Equipment:

   1. Atomic absorption spectrophotometer:   Any commercial  atomic absorption
      instrument having an energy source, an atomizer burner system,  a
      monochromator, and a detector is suitable
   2. Balance, analytical  with a 0.1 milligram sensitivity
   3. Hollow cathode lamps - cadmium, chromium, lead
   4. Hot plate, capable of holding at least ten 250 ml  beakers
   5. pH meter with single, combination electrode - optional for pH adjustment
   6. Pressure regulator valves:
      a. Two stage regulator designed to deliver acetylene  with an
         inlet CGA 510 connector
      b. Two stage regulator designed to deliver air with an inlet
         CGA 1340 connector
   7. Recorder:  One compatible with the electronics of the atomic
      absorption instrument is acceptable
   8. Steam bath for up to 9 - 100 ml beakers
      (sample, spike, duplicate, standards).
      determined,
   9, Still - borosilicate glass distillation apparatus  or  another source
      of good distilled water
  10. Stop watch
                               or 250 ml  beakers
                               Required if chromium is  to
B. Reusable Supplies:
   2.
   3.
      18 or 36
      (sample,
beakers, 9 or 13
spike, duplicate
             100
             - 4
             ml,
- 100 ml, 9 - 250 ml size, graduated,
, standards)
ml size, 2 brown glass, 4 clear
clear glass, 1000 ml capacity,
       clear glass, 100 ml
                                three
   Six dropper bottles -
   Eight Reagent bottles
   one brown glass, 1000
4. Cylinders - graduated
   1   500 ml
   1   250 ml
   2   100 ml
   1    25 or 50 ml
   1    10 ml
9 or 18 10 ml stoppered, wide base (sample, spike, duplicate, standard)
5. Flask, volumetric, glass stoppered
   1  1000 ml
   1   100 ml
6. Funnel, very small to filter 365 ml, glass
   1 (sample, spike, duplicate, standard)
7. Funnel, separatory, glass stoppered, teflon stopcock, 250 ml,
   1,9, or 18 (sample, spike, duplicate, standards)
8. Pi pets, graduated, mohr type
   1   5 ml
   2  10 ml
                                                                        E7-5

-------
     WATER MONITORING  PROCEDURE:   Determination of Total Cadmuim, Chromium and
                                  Lead  by Atomic Absorption

        8. Volumetric  type  (continued)
          4    1 ml
          2    2 ml
          1    3 ml
          4    5 ml
          5   10 ml
          1   20 ml
        9. Instrument  and manufacturer's operation manual
       10. Safety  glasses
       11. Separatory  funnel  rack
       12. Wash bottle,  plastic, squeeze type
       13. Watch glasses, 3 (sample,  spike,  duplicate),  3.5  inches  in  diameter


     C,  Consumable:
        1. Deionizing  column  - mixed  bed type
        2. Gases
          Fuel, acetylene  (C2H-)  - for use  with the  atomic  absorption Instrument,
          purified  grade,  380 Cf, CGA  510
          Oxidant,  air  - for use  with  the atomic  absorption instrument,
          dry grade,  2200  cf, CGA size 1340
        3. Filter  paper  - Whatman  #40
        4. Plastic weighing boats  - about  12
        5. Labels
        6. Marking pencil
        7. Reagents
          Ammonium  hydroxide
          Nitric  acid
          Hydrochloric  acid
          Cadmium sulfate  (3 CdSQ^SH-O)
          Chromium  trioxide
          Lead nitrate
          Potassium permanganate
          Pyrrolidine dithiocarbamic acid*
          Carbon  disulfide
          Chloroform
          Sodium  azide

           for pH  adjustment  - only if the  indicator is to be used:
               95% ethyl alcohol
               Bromophenol  blue


        *AvaiTable from Aldrich Chemical  Co.,  940  West St. Paul  Avenue,
         Milwaukee,  Wisconsin, 53233  (414/273-3850).
E7-6

-------
   HATER HUN1TQRING PROCEDURE:  Determination of Total Cadmium, Chromium and Lead by
                                Atomic Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A, Glassware Preparation
B. Sample Preservation
   and Handling
1. Wash In tap water with
   detergent and brush.
2. Rinse well with tap water.

3. Rinse with 1:1 nitric acid,

4. Rinse well with tap water.

5. Rinse with 1:1 hydrochloric
   acid.

6. Rinse well with tap water.

7. Rinse well with delonlzed
   distilled water.

B. If possible, reserve all
   glassware used in metal
   analyses for that purpose
   only.

1. Collect at least a 1 liter
   sample.


2. Add 5 ml of concentrated
   nitric acid per liter
   of sample.
                          3. The sample may be kept for
                             6 months before analysis.
la. Do not use chromic acid to clean this glassware.
Ib. This procedure also applies to sample containers,
Ic. Quality control checks may verify that some of
    these cleaning steps are not necessary.
2a. Remove all detergent.
                                                         8a. Contamination from other reagents is less likely
                                                             this way.
la. A quart sample container may be used.
Ib. The sample container should have been  cleaned
    using the procedure above.

2a. The pH must be less than 2.
2b. The nitric acid should be checked for  metals
    content before use.
2c. More acid may be necessary for samples with
    higher total dissolved solids.

3a. Good practice would dictate analysis of a sample
    as soon after collection as possible.
                                                                                                          E7-7

-------
   HATER MONITORING PROCEDURE:  Determination of Total Cadmium, Chromium and Lead by
           '                    Atomic Absorption
                                                                         E7-8
OPERATING PROCEDURES
      STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE NOTES
C. Reagent Preparation

   1. Delonized
      Distilled Water
   2. Nitric Add,
      Concentrated
   3. Nitric Acid (1:1)
   4. Hydrochloric Acid,
      Concentrated
Prepare approximately ten
(10) liters of deionlzed
distilled water.
No preparation is necessary
if metals of Interest are
absent.
Pour add into a 100 ml
dropper bottle.

Add SO ml of water to a
100 ml graduated cylinder.

Add 50 ml of concentrated
nitric acid to the same
graduated cylinder.

Allow to cool.
Transfer into tightly-
stoppered bottle for
storage.

Label bottle 1:1 nitric
acid.

No preparation is necessary
if metals of interest are
absent.
la. Prepare by passing distilled water through  a
    mixed bed of cation and anion exchange resins.
Ib. Use deionlzed distilled water for preparation
    of all standards, reagents  and dilutions  and
    also for the washing of equipment.

la. Run a reagent blank to check purity.   If  results
    show necessity, remove impurities by  distilling
    1:1 acid in an all glass (borosilicate) still. The
    redistilled acid (68.0%) is essentially as
    concentrated as non-redistilled (69.0 - 71.0%)
    acid.
la. Deionlzed distilled water.
                                                         2a.
                                                         2b.
                                                         2c.
                                                         2d.
                                                         2e.
la,
    Caution:   Do not reverse this  order of addition.
    Use safety glasses.
    Heat may  be generated.
    Prepare in a well-ventilated area.
    Larger amounts may be prepared.   Use equal
    amounts of water and acid.
Run a reagent blank to check purity.   If
necessary, remove impurities by distilling 1:1
acid in an all glass (borosilicate) still.  The
resulting redistilled acid is 20.2% HC1  (~6N)
in contrast to the usual  36.5-38% (-12N)
reagent grade acid.

-------
  MAItK HUHllUMNli HHUIUJUW,:   uecennuiaLiun ui
                                Atomic Absorption
DERATING  PROCEDURES
                              STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reagent Preparation
   (Continued)

   5. Hydrochloric Add
      (1:1)
   6. Hydrochloric Add
      (2.5* v/v)
7.  Ammonium Hydroxide
   (NH.OH)
   Concentrated
                       1.  Add 50 ml  of water to a
                          100 ml graduated cylinder.
                       2.  Add 50 ml  of concentrated
                          hydrochloric acid to the
                          same graduated cylinder.

                       3,  Allow to cool.

                       4.  Transfer Into tightly-
                          stoppered bottle for
                          storage.

                       5.  Label bottle 1:1 hydro-
                          chloric acid.

                       1.  Add about 50 ml  water to
                          a 100 ml volumetric flask.

                       2.  P1pet 2.5 ml concentrated
                          HC1 to the flask.
                          3. Cool and dilute to 100 ml
                             with water.

                          1. Pour the concentrated
                             NH.QH into a glass
                             dropper bottle.
la. If you have redistilled hydrochloric acid
    (C.4.1a), It is~-6N and this preparation 5
    Is not necessary.
Ib. This preparation requires a well-ventilated area,

2fl- Caution:  Do not reverse this order of addition.
2b. Use safety glasses, Heat may be generated.
2a. See C.4.1a..   If you have redistilled hydro-
    chloric acid, It Is a -6N.   Accordingly,
    you need 5 ml of redistilled acid for this  step.

3a. Store in dropper bottle (about 100 ml  vol.) This
    is used to adjust pH.

la. Use a hood to prevent Inhalation of fumes.  Avoid
    contact with  skin.   Wear protective equipment.
Ib. Only some drops of this are needed for the  pH
    adjustment of acidified samples.
Ic. A brown glass dropper bottle conserves the
    stability of  this reagent.
                                                                                                     E7-9

-------
HATER MONITORING PROCEDURE:
    Determination  of Total Cadmium, Chromium and Lead by
    Atomic Absorption
                                                                                                        E7-10
OPERATING PROCEDURES
         STEP  SEQUENCE
                    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE NOTES
C. Reagent Preparation
   (Continued)

   8.  Aninonium Hydroxide
      (NH^OH) 2N
   9.  Bromophenol  Blue
      Indicator
  10.  Pyrrolidine
      dithiocarbamic
      acid  (PDCA)  -
      chloroform
      solution
1. Dilute 13 ml
   trated NH.OH
   water.
of concen-
to 100 ml with
1. Dissolve 0.1 g of the solid
   in 100 ml of 50% ethyl
   alcohol.
1. Add 500 ml chloroform to
   a liter flask.
2. Add 18 ml of analytical
   grade pyrrolidine.
                         3. Add  15 ml of carbon
                            disulfide (CS^)  in small
                            portions with swirling.
                         4. Dilute  to  1 liter with
                            chloroform.
 la. This reagent should be prepared In the hood to
    prevent  inhalation of fumes.  Avoid contact
    with skin.  Wear protective equipment.
 lb. This Is  used to adjust pH.
 Ic. Use a brown dropper bottle to store.

 la. A platform balance can be used for weighing.
 lb. In order to prepare this solution 95% ethyl
    alcohol  should be diluted in half (i.e., 50 ml
    alcohol  to 50 ml water).
 Ic. This solution Is stable Indefinitely so long as
    It is kept In a tightly-stoppered dropper bottle
    to prevent evaporation.
 Id. This solution  is not necessary if a pM meter
    is used  for pH adjustments.

 la. This reagent should be prepared in a well ven-
    tilated  area (or hood)
 lb. Measure  500 ml with graduated cylinder.

2a. Pipet with a graduated plpet.
2b. Generates heat -- cool before proceeding.
2c. For supplier, see chemical list.
2d. CAUTION  -- reagent Is flammable,  toxic and
    corrosive.

 3a. Carbon disulfide is very odorous.   Prepare In
    hood or well  ventilated area.
 3b. Use a measuring pi pet.
 3c. CAUTION - Heat generated - cool  before proceeding

4a. This solution can be stored for several months
    if stored in a brown bottle in a  refrigerator.

-------
   MATER HUfllTORING PRDCEDURF:   Determination of Total Cadmium, Chromium and Lead by
                                Atomic Absorption
OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reagent Preparation
   (Continued)

   11. Potassium
       Permanganate
       Solution
       (KMn04) 0.1 N
   12. Sodium Azide
       Solution (NaNj
1.  Weigh out 3.20 grams of
   potassium permanganate.

2,  Add about 500 ml  of water
   to a 1000 ml volumetric
   flask.

3.  Transfer the potassium
   permanganate to the flask.

4.  Dilute to the mark.

5.  Mix thoroughly.

6.  Store in a tightly-stop-
   pered reagent bottle.


1.  Weigh out 100 mg sodium
   azide (NaN3).

2.  Add about 50 ml of water
   to a  100 ml volumetric
   flask.

3.  Transfer the sodium azide
   to the flask.

4.  Swirl to dissolve.

5.  Dilute to the mark.

6.  Stopper and mix thoroughly

7.  Store In a tightly-stop-
   pered bottle.
la.  Prepare 1f chromium 1s to be determined.
Ib.  Use a trip balance and a plastic weighing boat.
                                                          3a. Hash the weighing boat with water and add the
                                                              washings to the flask.
6a. Label with concentration and preparation date.
6b. When using,  transfer a portion to a 100 ml
    dropper bottle.

la. Prepare 1f chromium 1s to be determined.
                                                                                                     E7-11

-------
                                                                                                       E7-12
   >r/ATER MONITORING PROCEDURE:  Determination of Total  Cadmium, Chromium and Lead by
                                Atomic Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
 C.  Reagent  Preparation
    (Continued)

    13.  Stock  Cadmium
        Solution
   14. Stock Chromium
       Solution
 1. Weigh out 2.282 grams  of
   cadmium sulfate
   (3CdSOy8H20).

 2. Add about 500 ml of water
   to a 1000 ml volumetric
   flask.

 3. Transfer the cadmium sul-
   fate to the volumetric
   flask.

 4. Add 2.0 ml of concentrated
   nitric acid.

 5. Dilute to the mark.
6. Mix thoroughly.

7. Store in a tightly-stop-
   pered bottle,

1. Weigh out 1.923 grams of
   chromium trioxide (CrO_).

2. Add about 500 ml of water
   to a 1000 ml volumetric
   flask.
3. Transfer the chromium
   trioxide to the volumetric
   flask.
la. Use an analytical balance and a plastic weighing
    boat.
                                                          3a.  Use  a  wash bottle  and rinse  the weighing  boat
                                                              with water three times,adding  each wash to  the
                                                              flask.

                                                          4a.  Use a  5 ml  graduated  pi pet.
                                                         5a, The solution contains  1000 rag Cd/llter
                                                             (1 ml =  1 rng Cd).
7a.  Label with concentration and preparation date.
7b.  Store in a refrigerator.

la.  Use an analytical  balance and a plastic
    weighing boat.
                                                         3a. Use a wash bottle and rinse the weighing boat
                                                             with water three times,adding each wash to the
                                                             flask.

-------
WATER MONITORING PROCEDURE:  Determination of Total Cadmium, Chromium and Lead by
                             Atomic Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reagent Preparation
   (Continued)
  15. Stock Lead
      Solution
4» Add 2.0 ml of concentrated
   nitric acid.

5, Dilute to the mark.
                          6. Mix thoroughly.

                          7. Store in a tightly-stop-
                             pered reagent bottle
1. Weigh out 1.599 grams of
   lead nitrate (Pb(N03)2).

2. Add about 500 ml of water
   to a 1000 ml volumetric
   flask.

3. Transfer the lead nitrate
   to the volumetric flask.
                          4. Add 10  ml  of concentrated
                             nitric acid.

                          5. Dilute to the mark.
                          6.  Mix thoroughly.

                          7.  Store in a tightly-stop-
                             pered reagent bottle.
4a. Use a 5 ml graduated plpet.
                                                          5a. The solution contains 1000 mg Cr/liter
                                                              (1 ml = 1 mg Cr).
7a. Label with concentration and preparation
    date.
7b. Store In a refrigerator.

la. Use an analytical balance and a plastic weighing
    boat.
3a. Use a wash bottle and rinse the weighing boat
    with water three times adding each wash to the
    flask.

4a. Use a 10 ml graduated pipet.
                                5a.  With water.
                                5b.  The solution contains  1000 mg Pb/liter
                                    (1  ml « 1  mg Pb).
                                7a.  Label  with concentration  and  preparation  date.
                                7b.  Store  in  a refrigerator.
                                                                                                      E7-13

-------
    WATER MONITORING PROCEDURF:
         Determination of Total  Cadmium,  Chromium and  Lead,  by
         Atomic  Absorption
                                                                                                    E7-14
 OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D,  Instrument Set-up

   1.  Pre Warm-up
   Prepare the Instrument for
   Initial operation.
   2.  Lamp Installation
1.  Install  appropriate hollow
   cathode  lamp.
                          2. Align the lamp
                             intensity.
                  for maximum
                          3. Set appropriate wavelength,
la. Reference 1s made to the manufacturer's  manual  of
    operation.
Ib. Check power requirements and availability.
Ic. Provide adequate ventilation, including  a vent
    over instrument burner.
Id. Provide adequate space for instrument and work
    area,
le. Provide drain facility for the Instrument.

la. Hollow cathode lamps for lead, chromium  and
    cadmium must be available.
Ib. If the instrument 1s a single beam type, some
    method of warm-up for the hollow cathode lamps
    should be available.
Ic. Do not exceed the maximum current rating for  the
    lamps as this can seriously affect Its life
    and stability.
Id. Refer to the instrument  manufacturers manual  for
    proper Installation procedure.

2a. Check instrument manual  for proper procedure.
                                3a.  Pb,  283.3 nm;  Cd,  228.8  nmi Cr,  357.9 nm.

-------
                                H- •_>.«_ i iiiiiiu
                                                 tuiai
                                Atomic Absorption
                                      luruinium ana teaa oy
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
D,  Instrument Set-up
   (Continued)

   3.  Burner
      Optimization
1.  Install  the burner head.
                          2.  Attach the necessary gasses
                             to the Instrument.
                          3.  Align the burner to obtain
                             optimum absorption.
   4.  Check Aspiration
      Rate
1. Optimize the aspiration
   rate.
la. The usual burner head for direct aspiration Is
    the three slot Bollng head.  For aspiration of
    organic solvents a conventional head with a
    single slot 7.6 cm (3 Inches) is used.

2a. For this procedure, acetylene and air are used.
    Use purified grades of the gasses.
2b. Attach a pressure regulator to the tanks.  Use a
    CGA fitting of 510 for the acetylene and a 590
    or 1340 for the air.
2c. Connect cylinders through the regulator to the
    Inlet part of the Instrument with plastic
    pressure tubing,
2d. All cylinders should be securely fastened to
    prevent them from tipping over.

3a, The analysis of lead 1s exceptionally sensitive
    to turbulence and absorption bands in the flame.
    Therefore, some care should be taken to position
    the light beam In the most stable, center, portion
    of the flame.  To do this, first adjust the
    burner to maximize the absorbance reading with a
    lead standard.  Then aspirate a water blank and
    make minute adjustments 1n the burner alignment to
    minimize the signal.

la. Aspirate a standard Into the burner and adjust the
    aspiration rate until optimum absorbance is ob-
    tained.
                                                                                       V.D.3.2a
                                                                                       (P. 32)
                                                                                       VII.D.3.3a
                                                                                       (P- 33)
VII.D.4.la
(P. 33)
                                                                                                      E7-15

-------
   WATER MONITORING PROCEDURF:
      Determination of Total Cadmium, Chromium and Lead by
      Atomic Absorption
                                                                                                    E7-16
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E.  Sol utilization for
   "Total" Metals,
   IF NECESSARY
1.  Acidify the entire sample
   at  the time of collection
   with cone,  nitric  acid,
   5 ml/liter
                          2.  Proceed with the rest of
                             these steps only if
                             necessary.
                          3.  Transfer  200  ml  or  more
                             of  the  well-mixed sample
                             to  a  graduated  beaker
                             of  appropriate  size.
                          4.  Add  5  ml  1:1  hydrochloric
                             acid for  each 100  ml
                             of sample to  be  treated.
                          5. Heat  the  acidified  sample
                            In  the  beaker  on a  steam
                            bath  or a  hot  plate until
                            the volume  has been
                            reduced to  15-20 ml.  Make
                            certain the sample  does
                            not boil.

                          6. Remove  the  beaker and
                            allow contents to cool.
la. For metals other than Cd, Cr and Pb, consult
    the source of this procedure for possible
    modifications of this procedure.
Ib. The acid may have to be redistilled before use.
    See C.2. for details.

2a. For drinking water samples, this entire
    solubilization procedure Is necessary only
    If the samples contain visible suspended
    and/or settleable matter.

3a. 200 ml is a usual sample volume for metal
    concentrations less than 100 ixj/liter.   Choose
    a volume appropriate to the expected level of
    metal  concentration.  Cadmium and lead can be
    analyzed from the same sample aliquot.   A
    separate sample aliquot is recommended for
    chromium.
3b. Additional  volumes will be required to provide
    sufficient  final  volumes for additional  runs
    of the sample, e.g. as a duplicate, a spike,
    analysis for several elements, etc.

4a. IF THE SAMPLE IS  BEING PREPARED FOR FURNACE
    ANALYSIS, do  not  add the 1:1 hydrochloric acid.
4b. The acid may have to be distilled before
    use.  See C.4. and 5. for details.
                                                                                       VII.E.2.2a
                                                                                       (p.  33)
                                                                                       VII.E.3.3a
                                                                                       (p.  34)
                                                                                                                 VII.E.3.3b
                                                                                                                 (p.  35)

-------
   WAItK MONITORING PKULEDUKK:
       Determination  of Total  Cadmium,  Chromium and Lead  by
       Atomic  Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E.  Solubilization for
   "Total" Metals,
   IF NECESSARY
   (Continued)
7. Wash down the beaker
   walls with deionized
   distilled water.

8. If necessary, filter
   the sample Into a 250 ml
   separatory funnel If
   cadimium and/or lead Is
   to be determined.  Use
   a 250 ml graduated beaker
   as a receiver if chromium
   is to be determined.

9. Adjust the volume of the
   treated sample according
   to the requirements of
   the subsequent analytical
   procedure.
                         10. Continue with Procedure F,
                             Preparation of Standard
                             Dilutions
 7a.  Use a  small  volume of the water.
                                                          8a.  Filter If the sample Is turbid, If you see
                                                              particles or If past experience with the
                                                              sample source Indicates that you should.
                                                              (Filtration removes particles that could clog
                                                              the atomizer of an atomic absorption Instrument)
                                                          8b.  If filtration 1s not necessary, go to the next
                                                              step.
 9a.  If the sample Is  to be directly aspirated,  the
     final  volume may  be a reduction of the  original
     to effect up to a 10X concentration of  the
     sample.
 9b.  If the sample is  to be treated by a chelation -
     extraction procedure to determine Cd or Pb, the
     sample should be  quantitatively transferred to
     a 250  ml  separatory funnel  and brought  to the
     volume of the standards used to establish the
     standard  curve.  The final  volume used  in this
     write-up  Is 200 ml.
 9c.  To determine total  chromium, the sample should
     be brought to the volume to be used for the
     standards.  In this write-up,  the final  volume
     is 200 ml to be contained in a 250 ml beaker
     for use In Procedure G.

 9d.  If the sample is  to undergo furnace analysis,
     the treated sample  should be adjusted back  to
     the volume of the aliquot used for this
     Solubilization procedure.

lOa.  Calibration Standards must  be  analyzed  with the
     same procedures as  are applied to samples.

                                            __
                                                                                                                 I.E.9.9a
                                                                                                                 (p.  30)

-------
   WATER MONITORING PROCEDURE;
                                                                                                       E7-18
                                 Determination of Total  Cadmium,  Chromium and  Lead  by
                                 Atomic Absorption
OPERATING PROCEDURES
         STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
  TRAINING
GUIDE  NOTES
F. Preparation of
   Standard Dilutions
   1.  Primary Dilution
   2.  Intermediate
      Dilution
 1.  Add  about  500 ml of water
    to a 1000  ml volumetric
    flask.

 2.  Add  5.0 ml of concentrated
    nitric acid.

 3.  Pipet 10 ml of the stock
    solution(s) of the metal (s)
    of Interest into this
    volumetric flask.

 4.  Dilute to  the mark with
    water,

 5.  Mix  thoroughly and label.
1. Add about 100 ml of water
   to a 200 ml volumetric
   flask.

2. Add 1.0 ml of concentrated
   nitric acid.

3. Pipet 20 ml of the pri-
   mary dilution of the
   metal(s) of Interest Into
   the flask.

4. Dilute to the mark with
   water.

5. Mix thoroughly and label.
la. Estimate the amount.
2a. Use a 5 ml graduated plpet.
2b. Use safety glasses

3a. Use a 10 ml volumetric plpet for each measurement
3b. It saves time and glassware to prepare a mixture
    of the metals at this stage 1f more than one
    metal 1s of Interest.
5a. The solution contains 10 mg/liter of Cd and/or
    Cr and/or Pb.
5b. Ideally, this solution should be prepared at  the
    time of use.

la. Estimate the amount.
                                                         2a. Use a graduated pipet.
                                                         3a. Use a 20 ml volumetric pipet,
                                                         5a. The solution contains 1 mg/liter of Cd and/or
                                                             Cr and/or Pb.
I.F.
(P-  30)

-------
   WATER HUIIITORING PROCEDURE;
                           Determination  of Total  Cadmium,  Chromium and Lead by
                           Atomic  Absorption
OPERATING PROCEDURES
                            STEP SEQUENCE
                                    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                        TRAINING
                                                       GUIDE NOTES
F. Preparation of
   Standard Dilutions
   (Continued)
   3.
Calibration
Standards
1.  Prepare a blank and a
   series of calibration
   standards.
                                                    5b, This solution should be prepared at the time
                                                        of use.
la.  Prepare a blank and standards using volumetric
    pipets (1, 2, 5, 10, 20 ml) for measuring the
    intermediate dilution and a graduated cylinder
    for the water.
Ib.  To determine Cd and/or Pb, prepare one series of
    standards in six labeled, 250 ml  separatory
    funnels so they are ready for the extraction
    procedure (H).  Columns A and B in the Table on
    E7-20 are to be used to prepare 200 ml volumes.
    (Make sure the stopcock on each funnel 1s
    closed before you add solutions to 1t).
Ic.  To determine Cr, prepare a separate blank and
    series of standards in six labeled, 250 ml
    beakers so they are ready for the oxidation
    procedure (G).  The volumes in Columns A and B
    In the table E7-20 are to be used to prepare
    200 ml volumes.
Id.  Calibration standards should be prepared fresh
    for each run of samples.
le.  The standards are used to prepare a standard
    curve.
If.  Once the standard curve has been determined, it
    need not be redone each time the analysis is
    carried out.  However, it should be verified
    by running a blank and a calibration standard at
    the HCL.  Standards at the MCL are included in
    the Table on the next page:
    -For Cd, 2.0 ml Intermediate Dilution
    -For Cr, 10.0 ml Intermediate Dilution
    -For Pb, 10.0 ml Intermediate Dilution
                                                                                                                 VII.F.3.Lib
                                                                                                                 (P. 35)
                                                                                                                 VII.F.3.Lie
                                                                                                                 (P.  34)
                                                                                                          E7-19

-------
   WATER MONITORING PROCEDURF:
       Determination  of Total  Cadmium,  Chromium  and  Lead  by
       Atomic  Absorption
                                           E7-20
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                       TRAINING
                                      GUIDE  NOTES
F. Preparation of
   Standard Dilutions
   (Continued)
G. Oxidation for
   Total Chromium
                                Ig.  Table for preparing  standards:
                                                                     B
                                                               D
                          2. To determine Cr, continue
                             with Procedure G, oxida-
                             tion.  To determine Cd
                             and/or Pb, continue with
                             Procedure H, Extraction
                             of Metals.
1.  200 ml  volumes of sample(s)
   and of  standards should
   be in 250 ml  beakers  at
   this stage of the analysis.
                          2.  If necessary, adjust the
                             pH of each to 2.0 or less.

                          3.  Add 0.1 N potassium per-
                             manganate (KMnO.) drop-
                             wise to each solution
                             until a faint pink color
                             persists.
                          4.  Heat on a steam bath for
                             20 minutes,  adding
                                                         mrs of
                                                         Inter.
                                                         Diln.
                                1.0
                                2.0
                                5.0
                               10.0
                               20.0
                                        mi's  of
                                        Water
          200
          199
          198
          195
          190
          180
                 Cone.
                 (mg/1)
                 200 ml
      In
0.000
0.005
0.010
0.025
0.050
0.100
Cone.
(mg/1) in
Final  10 ml
   0.00
   0.10
   0.20
   0.50
   1.00
   2.00
Instrument
Reading
la. If It was not necessary to solublllze the sam-
    ple(s) (Procedure E), at this time:
    -measure 200 ml of each well-mixed sample into
     a labeled, 250 ml beaker.
    -measure 200 ml of one of the samples to run as
     a duplicate (in a 250 ml beaker).
    -measure 200 ml of one of the samples and spike
     It (in a 250 ml beaker).

2a. Use a pH meter to check pH.
2b. Use cone, nitric acid dropwlse to adjust the pH.

3a. The extraction procedure (H) will extract only
    hexavalent chromium.  To determine total
    chromium, you add potassium permanganate to
    oxidize any trlvalent chromium to the hexavalent
    species.
3b. If volume becomes a problem, use a more  con-
    centrated solution of KMnO..

4a. A slight excess of KMnO. must be maintained.
                                                       VII.F.3.1.1g
                                                       (p.  33)
                                     VII.G.I
                                     (p.  34)
                                     VII.G.I.la
                                     (p.  35)

-------
   WATtK HONITQRING PROCEDURE:
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
6. Oxidation for
   Total Chromium
   (Continued)
   additional drops of 0.1 N
   KMnO. to any solution in
   whicn the faint pink color
   disappears.

5. While still on the steam
   bath, add sodium azide
   solution (0.1%) to each
   solution dropwise until
   the faint pink color of
   the KMnO, Just disappears.

6. Continue heating the solu-
   tions for 5 minutes after
   adding the last drop of
   sodium azide solution.

7. Transfer the beakers to a
   water bath and cool to
   room temperature.

8. If necessary, filter the
   solution(s) Into a 250 ml
   separatory funnel(s).
                             If filtration was not
                             necessary, quantitatively
                             transfer the sample(s)
                             and standards to 250 ml
                             separatory funnels.
                                                          5a. Potassium permanganate can interfere with
                                                              subsequent processing.
                                                          5b. Heat for about 2 minutes after each addition
                                                              of sodium azide.  Avoid adding any excess.
8a. Use Whatman No. 40 or equivalent filter paper
    to filter any solution with a brownish preci-
    pitate or coloration which may Interfere with
    the pH adjustment 1n the extraction procedure
    (H).  If a pH meter is used in procedure H,
    you do not need to filter even if a solution
    is colored at this stage.

9a. Label each funnel to Identify the contents.
9b. Keep rinse volumes as small as possible during
    the transfer.
                                                                                       VII.G.9
                                                                                       (p. 35)
                                                                                                      E7-21

-------
   WATER MONITORING PROCEDURE:
                                                                           E7-22
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
H. Extraction of Metals
   1.  pH Adjustment
   200 ml volumes of sam-
   ple(s) and of standards
   should be in 250 ml
   separatory funnels at this
   stage of the analyses.
1. Use a pH meter and hydro-
   chloric acid (2.5* v/v)  to
   adjust the pH to 2.3 in
   each solution.  Then
   continue at H.2., Chela-
   tion and Extraction.

2. Add 2 drops of bromophenol
   blue Indicator to each
   sample and to each
   Standard.

3. Mix well.

4. Add ammonium hydroxide
   dropwlse until a very
   pale blue color persists.
                          5.  Add  2.5%  v/v  hydrochloric
                             acid dropwise until  the
                             blue color  just  disappears
la.  If it was not necessary to solubilize
    (Procedure E) or to oxidize (Procedure G)
    the sample(s), at this  time:
    -measure 200 ml  of each well-mixed  sample  into
     a labeled 250 ml separatory funnel.
    -measure 200 ml  of one  of the samples  to run
     as a duplicate (in a 250 ml  separatory funnel).
    -measure 200 ml  of one  of the samples  and  spike
     it (in a 250 ml  separatory funnel).

la.  A single combination electrode should  be used
    so the adjustment can be done in  the separatory
    funnels.
Ib.  If the pH meter and single combination electrode
    are not available, use  steps  2-6  to do the pH
    adjustment using bromophenol  blue indicator.
                                                          3a.  If  any  solution  Is pale  blue,  skip step 4.
                                                          4a.  Use  concentrated  NH.OH  for  acidified  samples  and
                                                              standards  made  with  10  or more ml  of  acidified
                                                              intermediate  dilution solution.  Use  2
                                                              for  more dilute standards.
                                                          4b.  The  reagents  should  be  in glass  dropper  bottles
                                                              for  this addition.   Use a hood.
                                           N  NH4OH
                                5a.  Use a glass dropper bottle for this  addition.
                                5b.  A pale yellow color may appear.
I.II.
(P.30)

VII.H.I
(P- 34)

VII.H.I.la
(p. 35)

-------
   WATER MONITORING PROCEDURE:
OPERATING PROCEDURES
H. Extraction of Metals
   (Continued)
   2, Chelation and
      Extraction
        STEP SEQUENCE
6. Add 2.0 ml 2.5% v/v
   hydrochloric acid.
   Stopper and shake.

1. Add 5.0 ml pyrrolidine
   dithiocarbamlc acid (POCA)
   reagent to the sample(s)
   and to each standard.
                          2. Shake each vigorously for
                             2 minutes
                          3. Allow the PDCA reagent to
                             settle to the bottom of
                             the separatory funnel.

                          4. Open the stopcock and
                             slowly drain off the lower
                             reagent phase of each into
                             a 100 ml beaker.

                          5. If total chromium is to be
                             extracted, re-adjust the
                             pH of the aqueous phases
                             in the separatory funnels
                             back to 2.3 before con-
                             tinuing.  Omit this step
                             if only Cd and/or Pb is
                             to be extracted.
    INFORMATION/OPERATING GOALS/SPEC IFICATIONS
6a. Use a 2.0 ml volumetric pipet.
6b. The pH at this point is 2,3
la. Use a 5.0 volumetric pipet for this step.
Ib. This reagent should be allowed to come to room
    temperature before pipetting, since it will be
    stored in a refrigerator,
Ic. The bottle should be restoppered immediately
    after use and returned to the refrigerator to
    prolong usefulness,

2a. CAUTION: Use proper technique with the separatory
    funnel.  The reagent contains volatile solvents
    and pressure Is formed which is released by
    opening the stopcock periodically.

3a. Enough time should be allowed for complete
    separation of the two phases.
3b. It may take up to 3 minutes.

4a. Hark each beaker with the number of ml used to
    prepare the standard or with the sample
    identification code.
                                5a. Use the steps in Procedure H.I., pH Adjustment
                                    (above), to adjust the pH back to 2.3 in each
                                    solution.
 TRAINING
GUIDE NOTES
                                                                                                     E7-23

-------
   HATER HUN1TORING PROCEDURE:
      Determination of Total Cadmium, Chromium and Lead by
      Atomic Absorption
                                                                                                      E7-24
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
H. Extraction of Metals
   (Continued)
   3.  Recovery of
      Complex
   4.  Digestion  of
      Complex
6. Add a second 5.0 ml of
   PDCA reagent to each
   separatory funnel.
7. Shake each vigorously for
   two minutes.

8. Allow the reagent to
   settle and separate.

9. Open the stopcock and
   slowly drain off the
   lower reagent phase into
   the beaker containing the
   reagent phase from the
   first 5.0 ml extraction
   of the sample or standard.

1. Evaporate each combined
   extract to dryness on a
   steam bath 1n a hood.
2, Remove and cool 2 minutes.

1. Add 2 ml concentrated
   nitric acid (HNOJ to
   each residue.
6a. The same volumetric pipet can be used for all
    additions of POCA to all samples and standards
    provided caution Is used to prevent contamina-
    tion.

7a. CAUTION: Use proper technique.  Open the stop-
    cock periodically to release pressure.

Ba. It should take about 2-4 minutes for complete
    separation of the two phases.

9a. A pale pink color may show In extracts.
la. The residue Is a light color with possible pale
    green or blue tinges.
Ib. Do not "bake" the residue,
Ic. Should take about 10-15 minutes.
la. Best carried out in a hood.   This is a violent
    reaction with boiling and dark brown fumes given
    off at the beginning.
Ib. Hold the beaker at a 45 degree angle.   Use a
    measuring pipet for the acid and add the acid
    down the walls, dropwise at  first while rotating
    the beaker.   When most of the residue  has
    dissolved, the acid can be added at a  faster  rate
Ic. The concentrated HNO., must be a good grade as any
    metals in the acid will be concentrated along
    with the sample.

-------
   HATER  MONITORING PROCEDURE:
       Determination  of  Total  Cadmium,  Chromium and  Lead by
       Atomic  Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
H. Extraction of Metals
   (Continued)
   5. Dissolving the
      Residue
2. Place the beakers on a
   low temperature hot plate
   or steam bath and eva-
   porate just up to dryness.

3. Remove from hot plate and
   cool for 2 minutes.

1. Add 2 ml of 1:1 nitric
   acid (HNOj to each
   beaker.

2. Return each to the low
   temperature hot plate or
   steam bath and heat for
   1 minute.

3. Cool and quantitatively
   transfer each solution to
   a labeled 10 ml volu-
   metric flask.
                          4. Bring each to the final
                             10,0 ml volume with
                             deionlzed distilled water.

                          5. Stopper each and mix well.

                          6. The sample(s) and standards
                             are now ready for aspira-
                             tion into the atomic
                             absorption instrument.
2a. Care should be taken to remove each beaker when
    only a very small amount of deep brown liquid
    remains in the beaker.
2b. The evaporation takes about 8 minutes.
la. Use a measuring pi pet.
Ib. Down inside walls at first.
                                                          2a. Both standards and samples should be carried
                                                              through this step at the same time as It could
                                                              affect the final concentration of add.
3a. A stirring rod and a plastic wash bottle con-
    taining deionlzed distilled water should be used
    to wash the beaker and transfer the solution.
3b. A wide base 10 ml volumetric flask Is suggested
    or place the volumetric flask in a beaker to
    prevent tipping it over.  A 10 ml stoppered
    graduated cylinder can be used instead of a
    volumetric flask.
3c. Mark the flask or cylinder with the number of ml
    used to prepare the standard or with the sample
    identification code.

4a. You might use a dropper to add the final amount
    of water.
                                                                                                     E7-25

-------
   HATER  MONITORING PROCEDURE:
      Determination of Total  Cadmium,  Chromium  and  Lead  by
      Atomic Absorption
                                                                           £7-26
OPERATING PROCEDURES
STEP SEQUENCE
                                    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE  NOTES
I.  Instrument
   Calibration
1. Turn on power to instru-
   ment and lamps.

2. Check and adjust to
   optimum settings all
   instrumental  operational
   parameters.
                          3.  Aspriate  the  blank  to
                             finalize  the  zero setting.

                          4.  After  the  standard  series
                             has  been  prepared (Section
                             F),  oxidized  (Section G)»
                             and  extracted (Section H),
                             begin  with the  lowest
                             standard  and  aspirate the
                             series Into the atomic
                             absorption instrument.

                          5.  Measure and record  the
                             peak height,  in milli-
                             meters, obtained on the
                             records.
                         6. Check all results before
                            proceeding.
la. If the power has been turned off.
2a. Set wavelength to one of the settings given
    earlier (D).
2b. Install proper lamp.
2c. Adjust current to the lamp as listed by
    manufacturer.
2d. Set slit width.
2e. Ignite flaw.
2f. Adjust fuel and oxidant flows to produce a
    blue flame.
2g. Adjust zero on recorder.
2h. See operational manual  for directions.

3a. After extraction.
                                4a.  If the  instrument  Is  to  be  adjusted to  read
                                    directly in  concentration It may be necessary to
                                    start with the  highest standard to set  the slope
                                    of the  absorbance.  The  instrument manufacturers
                                    manual  will  describe  the procedure.
                                Sa. The  various  instrument operational settings
                                   should  be  recorded  for the  record.
                                5b. Repeat  the aspiration of the standards and blank
                                   a  sufficient  number of times to secure a reliable
                                   average reading  for each.   The finalized readings
                                   could be recorded in column E on the Table in
                                   F.3.1.1g

-------
   HATER  HUHITORING PROCEDURE;
      Determination  of  Total  Cadmium,  Chromium  and  Lead by
      Atomic Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
I. Instrument
   Calibration
   (Continued)
7,  Plot the standard curve,
J, Calculations
K. Instrument Shut-Down
7a. Plot a curve for each metal.
7b. Plot on linear graph paper,
7c, Plot the peak height, 1n millimeters against the
    concentration, 1n mg metal  per liter, before
    extraction. - See Column C  on the Table, F,3,l.lg
7d. To check a standard curve,  run at least a blank
    and one standard at or near the MCL.  This check
    should be done with each sample or set of samples.
    The check should be within  +_ 10% of the original
    value.  If not, a new standard curve should be
    prepared.
                                                                                                                 VII.I.7.7d
                                                                                                                 (p. 35)
1, Read the metal concentra-
   1n each sample In mg/liter
   from the appropriate
   calibration curve (1.7).
la. So long as 200 ml portions of sample are used,
    the calibration curve can be used directly to
    obtain mg/Hter concentrations for samples.
Ib. If a sample was diluted to a 200 ml volume,
    multiply the curve reading by an appropriate
    dilution factor.
Ic. If multiples of 200 ml volumes of a sample were
    processed and extracts combined, multiply the
    curve reading by the appropriate factor.
                                                                                                                 Vll.J.l.lb
                                                                                                                 (p. 34 )

                                                                                                                 VII.J.l.lc
                                                                                                                 (p. 34)
1. If a flame Is burning.
   aspirate water for about
   15 seconds.

2. Close the acetylene
   cylinder valve.

3. Close the air cylinder
   valve.

4, Depress necessary switches
   to off.
la. This will prevent build-up of solids in the
    capillary.
                                                          2a.  The flame  will  automatically extinguish Itself,
                                                              leaving about 9 psig in the acetylene supply line.
                                                          4a. CAUTION: Exercise care In touching the burner
                                                              head and vent area.   These will  be hot enough
                                                              to cause serious burns.
                                                                                                    E7-27

-------
   l!ATF,n. MONITORING PROCEDURE:
      Determination  of  Total  Cadmium,  Chromium  and  Lead  by
      Atomic  Absorption
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
L. Maintenance
1.  Clean the instrument
   regularly.
                          2.  Insure the drain cup is
                             filled each day prior to
                             ignition.
la. A regular program of care and maintenance will
    prolong the life-time and maximize Its utility.
    Such items as filters in gas lines, air intakes,
    burner compartment, burner, and nebulizer should
    be cleaned.

2a. See the instrument manufacturer's manual  for
    exact procedures.
                                                                                                       E7-28

-------
                                  TRAINING GUIDE

 SECTION                         TOPIC
   I*                  Introduction
  II                   Educational Concepts - Hathematics
 HI                   Educational Concepts - Science
  IV                   Educational Concepts - Communications
   V*                  Field and Laboratory Equipment
  VI                   Field and Laboratory Reagents
 VII*                  Fteld and Laboratory Analysis
VIII                   Safety
  IX                   Records and Reports
  *Training guide materials are presented here under the headings marked*.
  These standardized headings are used through this series of procedures.
                                                                        E7-29

-------
WATER MONITORING PROCEDURES:
              Determination  of Total  Cadmium, Chromium  and  Lead by
              Atomic Absorption
INTRODUCTION
                                                           Section I
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
  F.
  H.
 E.9.9a
The National  Interim Primary Drinking Water Regu-
lations  listed ten inorganic parameters which are
required to be analyzed for by a public water
supplier.  Along with this list, a level was set for
the amount of that parameter permitted to be in the
water.  This was the "Maximum Contaminant Level or
MCL."  The MCL's for the metals included in this
procedure are:
                     Cadmium - 0.010 mg
                     Chromium - 0,05 mg
                     Lead - 0.05 rag per
                        per liter
                        per liter
                        liter
The range for calibration standards in this write-
up is based on the MCL concentrations.  The set
to be prepared will provide at least one standard
above and one below the HCL for cadmium or chromium
or lead, thus bracketing the concentration of
interest.

At such low levels as the MCL's an extraction pro-
cedure is usually the recommended method to con-
centrate samples so an instrument can detect the
metal.  Extraction levels recommended for these
three metals are:

     Cadmium - 0,020 mg per liter
     Chromium - 0.050 mg per liter
     Lead - 0.200 mg per liter

Accordingly, this write-up includes an extraction
procedure using PDCA - chloroform reagent.

Another way to concentrate metals in samples is to
evaporate a large volume of a sample at a low pH
and then directly aspirate.  Up to a 10X concen-
tration of a sample by evaporation is permissable
if these three conditions are met:

1. The total dissolved solids in the original sam-
   ple do not exceed 500 mg/liter.

2. The determination is corrected for non-specific
   absorbance.

3. There is no loss by precipitation.
                                 (CONTINUED)
                                                                      Methods for Chemical
                                                                      Analysis of Water and
                                                                      Wastes, 1979, EPA-EMSL,
                                                                      Cincinnati, Ohio 45268,
                                                                      Metals Section
                                                                      Ibid
  E7-30

-------
WATER MONITORING PROCEDURES;    Determination  of  Total  Cadmium,  Chromium and  Lead  by
                               Atomic  Absorption
 INTRODUCTION
    Section I
                                 TRAINING GUIDE NOTE
REFERENCES/RESOURCES
                 The results obtained by this short-cut should be
                 checked thoroughly before reporting.  The total
                 solids would cause light—scattering effects
                 producing high results.  The sample absorbance
                 should be checked at a non-absorbing wavelength.
                 If absorbance is still obtained at this wavelength,
                 then the scattering effect is contributing to the
                 sample value and the extraction procedure must be
                 carried out.

                 Sulfates tend to suppress the absorbance of energy
                 by lead causing low readings.  With a standard
                 concentration at the MCL accompanying all samples,
                 this interference could be detected.
                                                                                 £7-31

-------
MATER MONITORING PROCEDURES:   Determination  of Total  Cadmium,  Chromium and Lead by
                              Atomic  Absorption
Field and Laboratory Equipment
                                                          Section  V
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
0.3.2a
As acetylene (CHCH) is packed dissolved in acetone
(CH3COCH3)» cylinders should be stored only in an

upright position.  The acetone content of the gas
typically depends on the cylinder temperature and
pressure.  Avoid introducing acetone into the in-
strument.  Should this occur, the normal flame ob-
tained will have a slight pink tinge and yield an
abnormally high background signal.  To reduce
acetone carry-over, ft is desirable to allow
acetylene cylinders to stand undisturbed for at
least twenty-four (24) hours before use.  Replace
the cylinder when the cylinder reaches 50 psig.
Instrumentation Laboratoi
Incorporated.

Instrumentation Handbook
113 Hartwell  Avenue
Lexington, MA  02173
 E7-32

-------
WATER MONITORING PROCEDURES:
            Determination of Total Cadmium, Chromium and Lead by
            Atomic Absorption
Field and Laboratory Analysis
                                                         Section VII
                                 TRAINING GUIDE NOTE
                                                     REFERENCES/RESOURCES
D.3.3a
D.4.1a
F.3.1.1g
E.2.2a
After the extraction procedure in this analysis,
the volume of the calibration standards for any
of these metals is only 10 ml.  You will need
additional amounts of a high and low standard
to use in setting up the AA instrument.  Column D
in table F.3.1.1g. gives the range of the con-
centrations of the standards after extraction
when they are ready for aspiration into the
AA instrument.  You can prepare comparable
standards by further diluting the primary
dilution solution described in F.lTIFor
example, two nils of primary dilution further
diluted to a 200 ml volume has a concentra-
tion of 0.10 mg/liter.  (The 200 ml volume
should also contain 20 ml concentrated nitric
acid so the acid content is comparable to the
extracts).  Later, use the extracted blank and
standards for a final check.

If particulates are in the sample, Procedure E
1s carried out in order to solubiHze any of the
metal 1n the particles.  If particulates are
present and you do not carry out Procedure E,
filter the sample through a 0.45 micron filter,
extract, aspirate and report the value as
"dissolved" metal.
                                                                              E7-33

-------
WATER HONITOP.ING PROCEDURES:  Determination of Total Cadmium, Chromium and  Lead by
                             Atomic Absorption
Field and Laboratory Analysis
                                                                           Section VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
 E.3.3a
 F.3.I.1C
 E.3.3a
 G.I
 H.I
 J.l.lb
 J.I.Ic
 If chromium is the metal  of interest, the
 standards, samples, duplicate and spike should
 be treated as a totally separate determination.
 There is no problem in using a mixture of metals
 in the preparation of the chromium standards.
 However, problems can be caused if the reagents
 in the oxidation procedure for chromium introduce
 contamination in regard to metals other than
 chromium in the final solutions to be aspirated
 into the AA.  The alternative is to run blanks
 on the oxidation reagents along with "mixed"
 standards if you want to use one sample for all
 three of the metals.

 200 ml is a usual sample volume for the extraction
 procedure for metal concentrations less than
 100  ug/liter.  If use of 200 ml of sample gives
results too high to be on scale, consult the
source of procedure about doing a direct aspira-
tion of the sample.  Alternatively, a smaller
sample aliquot can be used and diluted to the
200 ml volume used for the standards to establish
a calibration curve for the extraction procedure.
The final calculation becomes:
     mg/liter metal  _
     in sample      ~
                                         l  C  '
                     where:
                     A»mg/l of metal in diluted sample  obtained
                       from a calibration curve
                     B=ml deionized distilled water used for
                       dilution

                     C=ml of sample aliquot
 E7-34
                 If use of a 200 ml sample gives too  low results,
                 more than one sample aliquot can be  treated and
                 the extracts combined"me final calculation
                 becomes:

                     mg/liter metal.   ,/B<
                     in sample     "    ^C~

                     where:
                     A=mg/l of metal in  sample obtained from a
                       calibration curve.

                     B=volume in ml of each of the standards' used
                       to develop the curve.
                     C=total volume in ml of sample  aliquots whose
                       extracts were combined.

-------
WATER MONITORING PROCEDURES:   Determination of Total Cadmium, Chromium and Lead by
                              Atomic Absorption
 Field and laboratry Analysis
                                                          Section VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
 E.3.3b
 G.I.la
 H.I.la
F.3.1.1b
G.9.
I.7.7d
 As  a  check  on  precision, you  should measure  a
 second  aliquot of  at  least  one  sample  1n  a set,
 and process it in  exactly the same manner as
 the first.   Ideally,  the results  from  the two
 aliquots  should agree within  the  range of two
 (three  maximum) standard deviations as established
 by  the  analyst.

 As  a  check  on  accuracy, you should prepare a
 spiked  aliquot of  a chosen  sample in a set
 and process it in  exactly the same manner as
 the unspiked sample  To determine the  amount of
 spike to  add,  you  must have some  knowledge of the
 concentration  of the unspiked sample either by
 analyzing it or from previous experience.  Add
 enough  spike to about double  a  concentration
 at  the  lower end of the standard  curve concen-
 trations.   If  the  sample is at  the intermediate
 part  of the curve, add sufficient spike to bring
 the concentration  to about  7S%  of the  curve
 range.  You can  use either  the  primary or the
 intermediate dilutions in Procedure F  for spiking,
 depending on the concentration you choose to add
 and keeping in  mind that added  volumes  should be
 minimal.  One ml of primary dilution adds 0.05
 mg/liter  in  a  200  ml sample volume.  One  ml of
 intermediate adds  0.005 rag/liter to a  200 nil
 volume.   The Table in F.3.1.1g.  can be a useful
 guide.  After  running the unspiked and spiked
 samples, calculate the % recovery.  Check the
 % recovery  for  the analysis as established by
 the analyst.

 Larger  volume  funnels may be  used.  Also, if this
 number  of separatory funnels  is not available,
 the standards can  be prepared and held in other
 type  containers  prior to the  actual extraction
 procedure.  Take care that the solutions do not
 get contaminated during the holding time.

When a  daily check of the standard curve falls
outside the W%  limit of the  procedure, reagents
prepared since the last check and any recently
 purchased chemicals should be checked before
doing  a new curve.   If possible, some old chemical
should be retained when a new batch is purchased.
Consequently, the new may be checked against  the
old, should doubt arise in its purity.
                                                                      Handbook for Analytical
                                                                      Quality Control,
                                                                      USEPA-AQCl, 1972,
                                                                      Cincinnati, Ohio 45268,
                                                                      Chapter 6
                                                                                 E7-35

-------
                         A PROTOTYPE  FOR  DEVELOPMENT  OF
                         ROUTINE  OPERATIONAL  PROCEDURES

                                     for  the

                  DETERMINATION OF MERCURY  USING THE  FLAMELESS
                    ATOMIC ABSORPTION (COLD VAPOR)  TECHNIQUE
                                  as  applied  in

                    WATER  AND WASTEWATER  TREATMENT  FACILITIES
                                   and  in the
                        MONITORING OF EFFLUENT WASTEWATERS
              National  Training and Operational  Technoloqy Center
                    Office of Water Program Operations
                   U.S. Environmental Protection Agency
CH.ME.hg.lab.WMPJ.il.77


                                                                          E8-1

-------
WATER MONITORING PROCEDURE:   Determination of Mercury Using the Flame!ess
                             Atomic Absorption (Cold Vapor) Technique


1. Analysis Objectives:

   The learner will  use  the  attached WMP to prepare a sample for analysis  in-
   cluding reagent and sample preparation.  A description of the instrumentation
   will  be presented.   However,  the learner should consult the manufacturer's
   direction for operation of any equipment.

2. Brief Description of  Analysis:

   This  procedure is a three step procedure which 1) chemically vaporizes  the
   sample, 2)  introduces the mercury and 3) determines the mercury by flameless
   atomic absorption techniques.

3. Applicability of  the  procedure:

   This  method is applicable to  drinking,  surface, and saline waters; domestic
   and industrial wastes.

   a.  Range of Concentration - The  method  is  recommended  for use in the  range
      of 0.2 wg Hg/liter using a  100 ml  sample;  the upper limit can be varied
      by instrument  expansion or  dilution.

   b.  Pretreatment of  the  Sample  -  No pretreatment is necessary as the chemical
      digestion procedure  inherent  in the  method  is sufficient.

   c.  Treatment of interferences  -  High  chlorides will  require additional permanga-
      nate (as much  as 25  ml).  Then additional  hydroxylamine sulfate is needed
      (25 ml)  and the  dead air space of  the sample container swept out by  air
      before addition  of the stannous sulfate.

      Some volatile  organics can  interfere but can be removed by a preliminary
      run without reagents.

   d.  Source of the  Method - Manual  of Methods for Chemical  Analyses of  Water
      and Wastes, 1974 ed.,  p. 119;  U.S. EPA  Technology Transfer,  Cinti.,  OH 45268.
                                                                      E8-3

-------
      WATER MONITORING PROCEDURE:   Determination of Mercury Using the Flameless
                                   Atomic Absorption (Cold Vapor) Technique
      Operating Procedures:
      A.  Equipment Preparation
      B.  Instrument Set-up
      C.  Reagent Preparation
      D.  Sample Handling and Preservation
      E.  Calibration
      F.  Sample Determination
      G.  Calculation
E8-4

-------
WATER MONITORING PROCEDURE:  Determination of Mercury Using the Flameless
                             Atomic Absorption (Cold Vapor) Technique


Equipment and Supply Requirements

A. Capital Equipment:

   1. Atomic absorption spectrophotometer - Any commercial atomic absorption
      instrument is suitable if it has an open burner head area in which to
      mount an absorption cell, and if it provides the sensitivity and
      stability for the analyses.  Also instruments designed specifically
      for the measurement of mercury using the cold vapor technique are
      commercially available and may be substituted,
   2. Mercury hollow cathode lamp
   3. Recorder - Any multi-range variable speed recorder that is compatible
      with the UV detection system is suitable.
   4. Absorption cell - See Figure 4.  The cell is constructed from glass
      or plexiglas tubing 25.4 mm O.D. x 114 mm (Note 1).  The ends are
      ground perpendicular to the longitudinal axis and quartz window
      (25.4 mm diameter x 1.6 mm thickness) are cemented in place.  Gas
      inlet and outlet ports (6.4 mm diameter) are attached approximately
      12 mm from each end.  The cell is strapped to a support and aligned
      in the light beam to give maximum transmittance,
   5. Analytical balance, 200 gram capacity
   6. Trip balance, 500 gram capacity
   7. Water bath, capable of maintaining 95°C temperature

B. Reusable Supplies;

   1. Air pump - Any peristaltic pump, with electronic speed control, capable
      of delivering 1 liter of air per minute may be used.  (Regulated com-
      pressed air can be used in an open one-pass system.)
   2. Six BOD bottles (plus one bottle is needed per sample)
   3. Volumetric flasks
      Six 1000 ml
      Four 100 ml
      One 250 ml
   4. Pi pets
      Five 10 ml graduated
      Two 1  ml graduated
      One 1  ml volumetric
      One 2 ml volumetric
      Three 10 ml volumetric
      One 5 ml volumetric
   5. One 100 ml graduated cylinder; two 25 ml graduated cylinders
   6. One Laboratory apron or coat
   7. One pair safety glasses
   8. One spatula
   9. One pipet bulb
  10. One wash bottle for distilled water
  11. One glass stirring rod (about 6 inches long)


Note 1:   An all glass absorption cell, 18 mm O.D. by 200 iron, with inlet 12 mm
from the end, 18 mm O.D, outlet in the center, and with quartz windows has
been found suitable.

-------
    WATER MONITORING PROCEDURE:  Determination of Mercury Using the Flameless
                                 Atomic Absorption (Cold Vapor) Technique

    Equipment and Supply Requirements  (Continued)

      12. One powder funnel
      13. Rubber stoppers - two size #2 (for drying tube)
      14. Fifteen feet of Tygon tubing
      15. One glass tubing - 6 inches  x 3/4 inch diameter
      16. One Rotometer  (any unit capable of measuring air flow of 1 liter/min.)
      17. One set cork hole borers
      18, One brush (for cleaning balance)

    The  following equipment is needed  depending on which method is chosen to trap
    the  mercury,

       1. Liquid trap

          a. Straight glass frit, coarse porosity, such as Corning #404260
          b. Filtering flask, such as  Corning #40058
          c. Rubber stopper, one hole  to accept frit
          d. Reagents, KMn04 and H2S04

       2. Solid trap

          a. Activated carbon such as  Barnebey and Cheney #580-13 or #580-22
             from:  Barnebey and Cheney
                    E. 8th Avenue & Cassidy Street
                    Columbus, OH 43219

                            or

                    Coleman Instruments
                    42 Madison St.
                    Maywood,  IL 60153
                    Item #50-160

          b. Glassware - Can be assembled similar to  the drying tube (Figure 3).

       3. Closed System

          The following  equipment is needed when using the closed system with a
          trap.

          a. Two position  valve, or stopcock, such as Corning  #442838
          b. Glass  "Y" shaped tubing connecter
          c. Pinch  clamp,  type used for stopping flow in tubing

    C. Consumable Supplies:

       1. Sulfuric  acid  (H?SO.) concentrated

       2. Nitric acid  (HN03) concentrated

       3. Potassium permanganate, KMnQ*

       4, Potassium persulfate, K0S000
                                 C. C O
       5. Sodium chloride, NaCl

E8-6

-------
WATER MONITORING PROCEDURE;   Determination of Mercury Using the Flameless
                             Atomic Absorption (Cold Vapor) Technique

Equipment and Supply Requirements (Continued)

   6. Hydroxylamine sulfate  (HONH2)-H2S04 or

      Hydroxylamine hydrochloride NFLOH-HCl
   7, Stannous sulfate, SnSO. or stannous chloride,  SnCl^

   8. Mercuric Chloride, HgCl2
   9. Hydrochloric acid concentrated
  10. Magnesium perchlorate, Mg(C10.)2 for drying tube,  20 g.
  11. Distilled water
  12. Sponges (for cleaning  laboratory table tops)
  13. Notebook for recording weights and readings
  14. Two pieces of glass tubing (5 mm diameter,  about two inches  long)  for
      the drying tube
  15. Glass wool (for drying tube)
  16. Plastic weighing boats (about 10)
  17. Pen or pencil
                                                                      Ffl_7

-------
     WATER MONITORING PROCEDURE:  Determination of Mercury Using  the  Flame!ess
                                  Atomic Absorption  (Cold Vapor)  Technique
             Sample
            Chemical Sample  Preparation

                 a. Oxidation of all mercury  to mercuric  form
                 b. Reduction of all mercuric mercury  to  metallic mercury
            Aeration
                 The metallic mercury  is  ciruculated  as  a  vapor  through
                 the system
             Flame!ess Atomic Absorption

                 Absorption of  energy  at  253.7  nm  from  a  hollow  cathode
                 lamp measured  by  a  photodetector
E8-8

-------
 WAFER MONITORING  PROCEDURE:
  Determination of Mercury Using the Flameless Atomic
  Absorption (Cold Vapor) Technique
 OPERATING PROCEDURES
A. Equipment Preparation

   1.  Cleaning of
      Glassware
   2.  Balance Preparation
   Instrumental  Set-up

   1.  Flow System
        STEP SEQUENCE
1. Wash with detergent.
2. Rinse with tap water.

3. Rinse with 1:1 nitric acid.


4. Rinse with tap water.

5. Rinse with 1:1 hydrochloric
   acid,

6, Rinse with tap water.

7. Rinse with distilled water.

1. Check all  balances for
   cleanliness and proper
   operation.
1.  Before operation of the
   instrument,  four additions
   to the system should be
   considered (Figure 1).
    lnFfiRMATION/OP£RATING GOALS/SPECIFICATIONS
la. Cleaning should be carried out in this order.
lb. Care should be taken to insure clean glassware,
Ic. If possible glassware should be reserved for
    mercury analysis only and separated from other
    glassware.
                                                           3a.  Add 500 nil  concentrated nitric acid (HNO,)  to
                                                               500 ml  distilled water.                 J
                                                           5a,  Add 500 ml  concentrated  hydrochloric  acid  (HC1)
                                                               to 500 ml  distilled water.
la.  There are two ways the flow system can be set up.
    It can be operated as a closed or open system.
    In the closed system the mercury vapor con-
    tinously passes  through the system until  wasted
    in the mercury trap by the operator.   In  the
    open system the  vapor passes through  the  absorp-
    tion tube only once and goes directly to  the trap.
    Which system is  chosen will dictate what  equipment
    is necessary.  Figure I shows the choices and the
    equipment necessary for each.
 TRAINING
GUIDE NOTES
                                                                                                       E8-9

-------
 WATER  MONITORING PROCEDURE:
   Determination of Mercury Using the Flameless Atomic
   Absorption  (Cold Vapor) Technique
                                                                                                         E8-10
 OPERATING PRCCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Instrumental  Set-up
   (Continued)
   2.  Mercury Trap -
      Liquid  Type
1. One of the following mer-
   cury traps should be in-
   cluded in the system.


2. For a liquid type trap,
   use a 250 ml side arm
   filtering flask.

3. Assemble as shown in
   Figure 2.

4. Insert straight gas  dis-
   persion tube or frit
   through the hole so  that
   the bottom or fritted end
   is about one inch above
   bottom of the flask.
                          5. Insert into filtering
                             flask.

                          6. Connect tygon tubing to top
                             end of frit and a second
                             piece of tygon tubing to
                             the side arm of filtering
                             flask.
                          7. Add 200 ml of 1:1 potassium
                             permanganate (KMnO.) -
                             sulfuric acid (H S04)

                             Reagent (Reagent #10).
la. Because of the toxic nature of mercury vapor,
    precaution must be taken to avoid contamination.
    The vapor will be held in the trap after it has
    been measured.

2a. Use a filtering flask such as Corning #400580
    or its equivalent.
                                                          3a. Use a #3 cork hole borer to make the hole.
                                                          4a. Frit should have a coarse porosity such as
                                                              Corning #404260 or equivalent.  The frit should
                                                              always fall below liquid level in the flask.
                                                              Should the level become low add more liquid
                                                              (Reagent #10).  The nonfritted end should be
                                                              lubricated and care taken when the frit is in-
                                                              serted through the stopper so as not to break
                                                              the frit and injure the worker.
                                6a.  Care  should  be  taken  so  that the  liquid  level  does
                                    not come  close  to  the opening of  the side arm  of
                                    the flask.   This could flood the  instrument if
                                    allowed to do so.   If flooding should occur,
                                    dismantle the absorption  tube and  clean  it and
                                    the tubing immediately.

                                7a.  A  solution of 0.25% iodine  in a 3%  potassium
                                    iodide  (KI)  solution  may  also be  used (Reagent 12)i
                                7b.  Filling the  flask  can be  postponed  until  all of
                                    the apparatus is assembled.

-------
WATER MONITORING PROCEDURE:
  Determination of Mercury Using the Flameless  Atomic
  Absorption (Cold Vapor)  Technique
 OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERA]ING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Instrumental  Set-up
   (Continued)

   2.  Mercury Trap -
      Solid Type
   3. Drying Tube
1. The apparatus can be pre-
   pared similar to the dry-
   ing tube (B-3) but packed
   with 2-3 grams of acti-
   vated carbon.

2, The equipment can be pur-
   chased with adsorbent as
   an option from the
   analyzer manufacturer.

1. Construct as shown in
   Figure 3.

2. Bore a hole through a
   number 2 stopper with a
   number 2 cork hole borer.
   Repeat with a second
   stopper.

3, Insert a 2 inch long piece
   of glass tubing (5 mm
   diameter) through each
   stopper allowing about 1/2
   inch protruding from each
   end.

4. Fill a 6 inch piece of 3/4
   inch diameter tubing with
   20 grams of magnesium
   perchlorate
la.  Locate after 2 position valve in closed system,
    Figure 1  (system two)  or after the analyzer in
    an open system, Figure 1 (system three).
                                                           2a.  Position  as  above.
la.  Place between sample container and instrument.
                                                           3a.  Care  should  be  taken  when  inserting  glass  tubing.
                                                           4a.  Other drying  agents  such  as  calcium  chloride
                                                               (Cad,)  may be  used.
                                                                                                       E8-11

-------
 WATER  MONITORING  PROCEDURE:
  Determination of Mercury Using the Flameless Atomic
  Absorption (Cold Vapor) Technique
                                                                                                         E8-12
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Instrumental  Set-up
   (Continued)
   4.  Rotometer
5. Use a small piece of glass
   wool in each end of the
   tube to prevent loss of
   granules.

6. Insert stopper prepared
   above in each end of tube.

7, Replace drying agent when
   needed.
1.  Must be capable of measur-
   ing a gas  flow of 1  liter
   per minute.
                             Connect one length of
                             tubing between the sample
                             container and the drying
                             tube through the rotometer
                             to the fitting of the
                             instrument.
                             A second length of tubing
                             should beqin at the out
                             fitting of the instrument
                             and proceed to the next
                             piece of equipment.
5a. The tube should not be packed so tight as to
    restrict gas flow.
7a. Replace magnesium perchlorate or any drying agent
    regularly.  These materials tend to cake and form
    a plug when their limit of saturation is ap-
    proached.  The length of time the material will
    last will vary with use and samples.  Experience
    will dictate a routine.

la. Place between water trap and instrument.  See
    Figure 1  for location.
Ib. The rotometer may be removed from the circuit
    after the instrument pump rate is checked.
lc. The flow rate should be checked periodically to
    insure flow rate has not changed.

2a. The connection must be made to the sample con-
    tainer by side arm.   Reverse tubing connections
    may flood the instrument with liquid.
                                3a.  See  Figure  1 for gas  flow path.

-------
 WATER  MONITORING  PROCEDURE:
   Determination of Mercury  Using  the  Flameless  Atomic
   Absorption (Cold Vapor) Technique
 OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Instrumental  Set-up
   (Continued)

   5. Two Position Valve
   6. Recorder
      (Optional
   7. Instrument
1.  A two position valve is
   necessary when using a
   closed system and a trap.
   Use stopcock, Corning No.
   442838 or equivalent for
   the two way valve or
   stopcock.

2.  One position of the valve
   should go through the
   trap to the sample con-
   tainer.  The other positior
   should by-pass the trap
   and be connected to the
   frit of the sample
   container.

1.  Any multi-range variable
   speed recorder that is
   compatible with the equip-
   ment is suitable.

1.  Follow instrument
   manufacturer's directions.
C. Reagent Preparation

   1. Sulfuric Acid 0.5 N
                                                           2a.  It is  important to maintain a  specific  air
                                                               volume in the system.   Once the system  is
                                                               calibrated,  this volume cannot be  changed  unless
                                                               the system is recalibrated.
la.  Use of a recorder or its equivalent for analysis
    of potable water samples is strongly recommended,
   Add 14.0 ml  concentrated
   sulfuric acid (H2S04)  to

   approximately 500 ml  water
   and mix.  Then dilute with
   water to 1  liter volume.
la.  The concentrated H-SO.  should be of low mercury
    concentration.
Ib.  Unless specified the term water means  distilled
    water.
Ic.  Use a 25 ml  graduated cylinder to measure the
    sulfuric acid.
                                                                                                       E8-13

-------
 HATER MONITORING  PROCEDURE:
                           Determination of Mercury Using the Flameless Atomic
                           Absorption  (Cold Vapor) Technique
                                                                                                        E8-14
 OPERATING PROCEDURES
                                STEP SEQUEhCt
                                    INFORMATION/OPERA! ING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE  NOTES
C.  Reagent Preparation
   (Continued)

   2.  Potassium
      Permanganate
      Solution  -  5%
      Solution  w/v
   3.  Potassium Persul-
      fate  Solution  -
      5% Solution  w/v
4, Sodium Chloride -
   Hydroxylamine
   Sulfate Solution
   Solution
   122 NaCl
   12%
   ((HQNH2)2-H2SO
5.
      Stannous Sulfate
      Solution -  10%
      Solution w/v
6.
     Sulfuric Acid
     Concentrated
     (H2S04)
                        1. Prepare 100 ml of solution
                           containing 5.0 grams
                           potassium permanganate
                           (KMn04).
                        1. Dissolve 5.0 g potassium
                           persulfate (K0S00Q) in
                                        f. L O
                           water and dilute to 100 ml
1. Dissolve 12.0 g of sodium
   chloride (NaCl) and 12.0 g
   of hydroxylamine sulfate
          K-H-SO,) in water
   and dilute to 100 ml.
1. Add 25.0 g Stannous sul-
   fate (SnSOj  to about 40

   0.5 N sulfuric acid and
   dilute with 0.5 sulfuric
   acid to 50 ml.
1.  No preparation necessary.
                                la. Should a larger amount of reagent solution be
                                    needed the same ratio should be maintained.  For
                                    example:  prepare 1000 ml of solution containing
                                    50 grams KMn04-

                                Ib. Weigh out in plastic weighing boat on a trip
                                    balance.

                                la. Weigh out in a plastic weighing boat on a trip
                                    balance.
                                                          la. Hydroxylamine hydrochloride (NH~OH-C1) may also

                                                              be used.  It should be prepared in the same
                                                              manner.
                                                          Ib. Weigh out in a plastic weighing boat on a trip
                                                              balance.
la. Stannous chloride (SnCl2'2H?0)  may be used and be

    prepared in the same manner.
Ib. The acid is reagent no.  1.
Ic. This is a suspension and should be stirred con-
    tinuously during use.
Id. Weigh out in a plastic weighing boat on a trip
    balance.

la. This should be reagent grade  and low in mercury
    concentration.
Ib. Caution:  this is corrosive.

-------
WATER MONITORING PROCEDURE;
                        Determination of Mercury Using the Flameless Atomic
                        Absorption (Cold Vapor) Technique
 OPERATING PROCEDURES
                             STTP SEQUENCE
                                    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                        TRAINING
                                                       GUIDE  NOTES
C.  Reagent Preparation
   (Continued)
   7.
   8.
Nitric Acid
Concentrated
(HN03)
Stock Mercury
Solution (HgCl.
   9.
Intermediate
Mercury Solution -
      (HgCl,
          Dilution
      of Solution 8
1. No preparation necessary.
1. Dissolve 0.1354 g of mer-
   curic chloride (HgCl?) in
   water.
                           2.  Add 10 ml  concentrated
                              nitric acid
                           3.  Cool  to room temperature.

                           4.  Dilute to 100 ml  with
                              water.
1. The intermediate solution
   is a dilution of the stock
   solution to adjust the
   concentration of Hg to
   0.1 Kg/ml.   Proceed as
   follows.

2. Add about 700 ml water to
   a 1000 ml volumetric flask

3. Add 0.5 ml  concentrated
la.  This should be reagent grade and low in mercury
    content.
lb-  Ca_uti_gn_:   this is corrosive.
Ic.  If a high reagent blank is obtained, it may be
    necessary to distill  the nitric acid.

la.  Weigh out in a plastic weighing boat on an
    analytical  balance.
                                                     2a.  Caution:   solution will  increase in temperature.
                                                     2b.  Use a 10  ml  graduated pipet.
4a.  Concentration of stock solution is now
    1  ml  = 1  mg Hg.
4b.  Stock solution is stable for several  months.

la.  Prepare fresh before use.
                                                           3a.  The  nitric  acid  concentration  of  the  dilutions
                                                               including the  working  solution  should  be  main-
                                                               tained  at 0.15%.   This  acid  should  be  added  to
                                                               the  flask before addition  of the  aliquot.
                                                           3b.  Use  a  1 ml  pipet graduated in  tenths.
                                                                                                       E8-15

-------
 WATER MONITORING  PROCEDURE;
                              Determination of Mercury Using the Flameless Atomic
                              Absorption (Cold Vapor) Technique
                                                                                                        E8-16
 OPERATING PROCEDURES
                                   STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                                                         TRAINING
                                                                                       GUIDE  NOTES
C. Reagent Preparation
   (Continued)
  10.  Working Mercury
      Solution (HgCl2)

      Dilution of
      Solution 9
Prepare 11  or 12 for a
liquid trap.

  11.  Potassium
      Permanganate 0.1
      (KMn04) and

      Sulfuric Acid 10%
      Solution (for
      Mercury Trap)
                       M
                           4. Add 10 ml  stock Hg
                              solution.

                           5. Dilute to  1000 ml  mark.
                              This solution contains
                              10
                           1 .  Add about 700 ml  water to
                              a 1000 ml volumetric flask

                           2.  Add 1.5 ml  concentrated
                              HN03.

                           3.  Add 10 ml of intermediate
                              solution (10 yg/ml ) .

                           4.  Dilute to 1000 ml  mark.
                              This is working solution
                              and contains 0.1  yg/ml .
1. Dissolve .316 g potassium
   permanganate (KMnO.) in

   100 ml water.

2. Add 10 ml  concentrated
   sulfuric acid (H^SO.) to

   about 80 ml  water.   Dilute
   to 100 ml  with water.

3. Mix equal  volumes of each
   solution;  KMN04 (1) and
     Hen  /9\
    oOU. \l).
4a. Use a 10 ml volumetric pi pet.
2a. Use a 10 ml graduated pipet.
                                                           3a,  Use  a  10 ml  volumetric  pipet.
la. Let stand until following solution is prepared.
Ib. Weigh out in a plastic weighing boat or a trip
    balance.


2a. Caution:   heat generated.
2b. Should be at room temperature before volume
    adjustment.
2c. Use a 10 ml  graduated pipet.

-------
WATER MONITORING PROCEDURE:
   Determination of Mercury Using the Flameless  Atomic
   Absorption (Cold Vapor)  Technique
 OPERATING PROCEDURES
        STEP  SEQUENCE
   INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C.  Reagent Preparation
   (Continued)

  12.  Iodine (0.25) in
      KI (3%) Solution
D. Sample Handling and
   Preservation
1.  Measure 250 ml  of dis-
   tilled water in a gradu-
   ated cylinder.

2.  Transfer about  half the
   water to a 500  ml
   Erlenmeyer flask.

3.  Weigh out 7.5 grams of
   potassium iodide (KI).

4.  Transfer the potassium
   iodide to the 500 ml
   Erlenmeyer flask and
   dissolve.

5.  Weigh out 0.63  grams of
   iodine (I).

6.  Add to the Erlenmeyer
   flask and dissolve.

7.  Add the remainder of the
   water.

8.  Mix well.
                                                           3a.  A trip balance can be used,
                                                           5a.  A trip balance can be used,
1. Upon collection, the sample
   pH should be lowered to 2
   or lower by the addition
   of concentrated nitric
   acid (HN03).
la.  If only dissolved mercury is to be determined,
    the sample should be filtered before addition
    of the acid.   For total  mercury the filtration
    is omitted.
Ib.  If nitric acid cannot be used because of shipping
    restrictions, the sample may be initially pre-
    served by icing and immediately shipped to the
                  (Continued)
                                                                                                      E8-17

-------
 WATER MONITORING PROCEDURE:
    Determination of Mercury Using the Flameless Atomic
    Absorption (Cold Vapor) Technique
                                                                                                        E8-18
 OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE  NOTES
D. Sample Handling and
   Preservation
E.  Calibration
                           2. Collect samples in
                              appropriate container.

                           3. Do not store samples beyonc
                              the acceptable time.
                                    laboratory.  Upon receipt in the laboratory, the
                                    sample must be acidified with concentrated nitric
                                    acid to pH<2.

                                2a. Samples should be collected in acid-washed glass
                                    or high density polyethylene bottles.

                                3a. Samples should be analyzed within 38 days if
                                    collected in glass bottles and within  13 days
                                    if collected in polyethylene bottles.
1.  Prepare a series  of
   standards from the working
   mercury standard.
                                                           la.  Prepare as  follows:
                                                               ml 's  of
                                                               Standard
                                                                 0.0
                                                                 0.5
                                                                 1.0
                                                                 2.0
                                                                   0
                                                                10.0
                   ml's of Water

                       100.0
                        99.5
                        99.0
                        98.0
                        95.0
                        90.0
                                                                       Concentrated
                                                                          0.00
                                                                          0.05
                                                                          0.10
                                                                          0.20
                                                                          0.50
                                                                          1.00
                          2. Mix thoroughly.
                          3. Add 5 ml concentrated
                             sulfuric aci
                             each bottle.
  sulfuric acid (H9SO.) to
                                                           Ib.  Use 300 ml  BOD bottles.
                                                           Ic.  Use volumetric pipets  to pipet the standards.
                                                           Id.  A 100 ml graduated cylinder may be used for the
                                                               water.
3a. Mix thoroughly.
3b. Use caution when using concentrated acids
3c. Use a 5 ml graduated pipet.
I.E.3
VII.E.3
(p. 22 &
P. 26)

-------
 WATER  MONITORING  PROCEDURE:
Determination of Mercury Using the Flameless Atomic
Absorption (Cold Vapor) Technique
 OPERATING PROCEDURES
E.  Calibration
   (Continued)
     STEP S
4. Add 2.5 ml concentrated
   nitric acid (HNO,) to

   each bottle.

5. Add 15 ml potassium
   permanganate (KMnO*)
   (Reagent no.  2).


5. Allow to stand 15 minutes.

7. Add 8 ml potassium
   persulfate (K^S^On)

   (Reagent no.  3).

8. Heat in a water bath at
   95°C for 2 hours.
 9.  Cool  to room temperature.

10.  Add 6 ml  of the sodium
    chloride-hydroxylamine
    sulfate (NaCl-(HONH2)2-

    HpSO.) (Reagent no.  4).


11.  Allow to  stand at least
    30 seconds.
                                    inFURMATlON/OPERATING GOALS/SPECIFICATIONS
                             4a.  Mix thoroughly.
                             4b.  Use caution when using concentrated acids.
                             4c.  Use a 5 ml  graduated pipet.

                             5a.  Shake and add additional  potassium permanganate
                                 solution, if necessary, until  the purple color
                                 persists for the following standing time.
                             5b.  Use a 25 ml graduated cylinder for the addition,
                                                           7a.  Use  a  10 ml  graduated  pipet.
                                                           8a.  The  heat  step  is  required  for  organic  mercury
                                                               compounds.   For  standards  prepared with  dis-
                                                               tilled  water and  mercuric  chloride,  the  heating
                                                               step is not  necessary.
                                                          lOa.
                                                          lOb.
                                                          lOc.
                                                          lla
                                 Shake we!1.
                                 Use a 10 ml  graduated pipet.
                                 This reagent should decolorize the solution.
                                 6 ml is not  enough, add sufficient extra  to
                                 complete the decolorization.
                                                                                  If
                                 Up to this point all  samples  to be run can be
                                 treated as a group.   From this  point each must be
                                 done individually as  the mercury is liberated
                                 immediately upon addition of  the stannous
                                 sulfate.
                                                                                         TRAINING
                                                                                        GUIDE NOTES
                                                                                                         E8-19

-------
 WATER MONITORING PROCEDURE;
    Determination of Mercury Using the Flameless  Atomic
    Absorption (Cold Vapor)  Technique
                                                                                                        E8-20
  OPERATING PRCCEDURES
         STEP  SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE NOTES
 E.  Calibration
    (Continued)
12.  Add 5 ml  of stannous
    sulfate suspension  (SnSO.)
    (Reagent no.  5).

13.  Immediately insert  bubbler
    into the bottle.
                          14. Observe reading.
                          15. Trap the mercury.
                          16.  Continue with the
                              remaining standards.

                          17.  Plot a standard curve.
12a. Use a 5 ml graduated pipet with a large bore
     opening.
13a. The air mover (pump-air, cyl inder, etc.) should be
     calibrated before use at 1  liter/minute.  It
     should be on and allowed to  operate continously
     when analyzing samples or standards.
13b. The reading will  increase and reach a maximum
     within one minute.

14a. When a recorder is  used, allow the pen to level
     off.

15a. In the closed system the valve must be turned to
     the trap position until  the  absorbance returns to
     the minimum value.   The aeration should be
     continued.  In the  open system wait until  the
     absorbance returns  to the minimum value.

16a. Repeat steps 12-15. Section  D.
                               17a. Plot the peak height (from the recorder plot)
                                    versus the concentration in micrograms of
                                    mercury on arithmetic paper.
                                                                                       VII.E.15
                                                                                       (P.  26)
                                                                                                                  II.E.17a
                                                                                                                  (p.  24)
F.  Sample Determination
1. Transfer 100 ml of the
   sample to a 300 ml BOD
   bottle.
 la.
                                                           Ib.
The sample should not contain more mercury than
the upper limit of the equipment being used.  The
range of the method may be varied through
instrument and/or recorder expansion and by using
a larger volume of sample.
The usual range of the procedure is 0.2 to about
10.0 yg Hg/liter (.02 to 1.0 ugHg/100 ml).

-------
WATER MONITORING PROCEDURE:
                              Determination of Mercury Using the Flameless Atomic
                              At,sorption (cold Vapor) Technique
 OPERATING PROCEDURES
F. Sample Determination
   (Continued)
G. Calculations
                          2. Treat sample(s) with same
                             procedure used in cali-
                             bration section steps
                             3 through 15.
                          1. Determine the peak height
                             of the unknown(s) from the
                             recorder chart and read the
                             mercury value from the
                             standard curve,

                          2, Calculate the mercury con-
                             centration in the sample.
                           3.  Report mercury concentra-
                              tion as follows:

                              Below 0.2 yg/1 as <0.2 yg/1
                              Between 1.0  and 10.0 yg/1
                                using one  place after
                                decimal
                              Above 10.0 yg/1 using only
                                whole numbers
                                                              !NFC
-------
    WATER MONITORING PROCEDURE:   Determination of Mercury Using the nameless
                                 Atomic Absorption (Cold Vapor) Technique
                                   TRAINING GUIDE
    SECTION                           TOPIC
       I*                     Introduction
      II*                     Educational Concepts - Mathematics
     III*                     Educational Concepts - Science
      IV                      Educational Concepts - Communications
       V*                     Field and Laboratory Equipment
      VI                      Field and Laboratory Reagents
     VII*                     Field and Laboratory Analysis
    VIII                      Safety
      IX                      Records and Reports
     *Training Guide materials are presented here under the headings  marked*.
     These standardized headings are used through this  series  of procedures.
E8-22

-------
WATER MONITORING PROCEDURES:
              Determination of Mercury Using the Flameless
              Atomic Absorption (Cold Vapor) Technique
INTRODUCTION
                                                           Section I
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
E.3
Theoretical Concepts
                The method is based on the method developed by Hatch
                and Ott.   The mercury in the sample is oxidized to
                the mercuric ion with potassium permanganate in a
                nitric-sulfuric acid medium.   Hydroxylamine sulfate
                is  then  added to remove the excess permanganate,
                Stannous  sulfate,  Is then added to reduce the
                mercury  to metallic form.   Then the mercury is
                vaporized and circulated by the bubbler system.   Thi
                consists  of a circulating pump and the bubbler.
                Measurement is made with a flameless  atomic ab-
                sorption  spectrophotometer.   The energy at the 253,7
                nm  mercury line emitted by a  mercury  lamp is ab-
                sorbed by the mercury vapor in the flow-through
                absorption cell.   The change  in transmittance is
                detected  by a phototube.
                                                      Analytical Chemistry,
                                                      vol.  40, p. 2085,
                                                      December 1968.
                                                                              E8-23

-------
HATER MONITORING PROCEDURES:   Determination  of  Mercury  Using  the  Flameless
"                              Atomic  Absorption (Cold Vapor]  Technique
 EDUCATIONAL  CONCEPTS  -  MATHEMATICS
                                                           Section II
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 E.17a
The standard curve is a reference to a fundamental
law of absorption chemistry known as the Beer-
Lambert law.  Simply, this law states that the amount
of energy absorbed by a solution is proportional to
the concentration of the absorbing material in the
solution.  Applied to this outline the amount of
energy absorbed at the wavelength of 253.7 nm Is
proportional to the amount of mercury present in a
solution.

If the concentrations of a series of known solutions
(prepared as in step 1 under Calibration) and peak
height are plotted, a straight line should result.

When an unknown sample value is obtained, its mercury
content can be determined from the straight line or
standard curve.

To use the attached graph paper prepare the standards
as in step 1 of the Calibration section and run them
as described.   The known concentrations are plotted
along the bottom of the graph.   Then plot the peak
heights obtained from the recorder chart plotted on
the left of the graph.

Where the known concentration line intersects with
the appropriate peak height,  a  mark is  made.   After
all six standards are plotted,  draw a line through
the marks.
  E8-24

-------
WATER MONITORING PROCEDURES;
              Determination of Mercury Using the Flameless
              Atomic Absorption (Cold Vapor) Technique
FIELD AND LABORATORY EQUIPMENT
                                                           Section V
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
B.6.1a
The criteria for certification of a laboratory for
analysis of drinking water samples for compliance
rfith the Safe Drinking Water Act (93-523) lists as
nandatory equipment a recorder to be used when
inalyzing for mercury.
                                                                              E8-25

-------
WATER MONITORING PROCEDURES:
              Determination of Mercury Using the Flameless
              Atomic Absorption (Cold Vapor] Technique
 FIELD  AND  LABORATORY  ANALYSIS
                                                           Section VII
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 E.3
 E.15
Loss of mercury may occur at elevated temperature.
However, with the stated amounts of acid the
temperature rise is only about 13°C (25-38°C) and no
losses of mercury will occur.
 PA Method Study B» Total
lercury in Water,
 PA-600/4-77/012, Feb.  1977,
Be sure to close the valve before running another
sample/standard when using the closed system.  Since
the system is being calibrated with the valve in
that position, running a sample/standard in the open
mode will  produce a nonacceptable value.  This is due
to the change in volume of the system.

Operation  in the open mode does not require the
valve.  Consequently, the opening and closing will
not have to be done.
   E8-26

-------
.7
.61
.5;
.4-;
 .2
 .1
             10
20
3d
40          50
     Peak Height
60
70
100

-------
                            RECORDER
                          o
                    MERCURY
                     TRAP
                                          ROTOMETER
                   SAMPLE
                   CONTAINER
                   (BOD BOTTLE)
                   SYSTEM ONE: LIQUID MERCURY TRAP CLOSED SYSTEM
                           RECORDER
                      -o
                                WATER   UROTOMETER
                                 TRAP
                      SAMPLE
                      CONTAINER
                      (BOD iOTTLE)
                   SYSTEM TWO: SOLID MERCURY TRAP CLOSED SYSTEM
                      SAMPLE  CONTAINER
                      (BOD BOTTLE;
                 SYSTEM THREE: SOLID MERCURY TRAP OPEN SYSTEM
Es-28
Figure  1. FLOW  SYSTEMS FOR THE COLD VAPOR
          TECHNIQUE  FOP MFPrilRY

-------
                                                         FROM
                                                        MAS-50
                       RUBBER STOPPER
                             (size #6)
                  FILTERING FLASK
                  (Corning #400580
                                                                 TYGON TUBING
                                                                  STOPCOCK
                                                                     GLASS "Y" SHAPED
                                                                     TUBING CONNECTOR
TO SAMPLE
CONTAINER
                                         SIDE ARM OF
                                         FILTERING FLASK
CD
I
ro
id
                   FIGURE 2. ARRANGEMENT OF TWO-POSITION STOPCOCK AND
                              MERCURY TRAP

-------
oo
i
U)
o
                    GLASS

                    TUBING
6 inches
                                  MAGNESIUM PERCHLORATE (20 g)
                    RUBBER STOPPER     GLASS WOOL

                    (#2 size)
                                 3/4 inch

                   __^	 ^_ 1  Jf_ diameter




           GLASS WOOL     TYGON

                           TUBING
                                     FIGURE 3. DRYING TUBE

-------
                                                   18 MM
                                                   18 MM
                                                   JL
                            10 - 18 CM
       FIGURE  4.CELL FOR MERCURY  MEASUREMENT
                   BY  COLD VAPOR  TECHNIQUE
The length and OD of the cell are not critical.   The body of the cell
may be of any tubular material but the end windows must be of quartz
because of the need for UV transparency.

The length and diameter of the inlet and outlet  tubes are not important,
but the position of the side arms may be a factor in eliminating recorder
noise.  There is some evidence that displacement of the air inlet arm
away from the end of the cell results in smoother readings,  A mild
pressure in the cell can be tolerated, but too much pressure may cause
the glued-on end windows to pop off.

Cells of this type may be purchased from various supply houses.
                                                               E8-31

-------
                               A PROTOTYPE FOR DEVELOPMENT OF
                               ROUTINE OPERATIONAL PROCEDURES

                                          for the

                           DETERMINATION OF ARSENIC AND SELENIUM
                                       as  applied in

                            DRINKING WATER TREATMENT FACILITIES
                                         and in the
               DISTRIBUTION SYSTEMS OF DRINKING WATER TREATMENT FACILITIES
                  National Training and Operational Technology Center
                           Office  of  Water  Program Operations
                           U.S.  Environmental  Protection  Agency
CH.ME.lab.WMP.2.11.77

                                                                               E9-1

-------
WATER MONITORING PROCEDURE:  Determination of Arsenic and Selenium


General Description of Equipment and Supplies Used in the Process

A.  Capital Equipment:

    1.  Balance, analytical - sensitivity 0.1 milligrams
    2.  Atomic absorption spectrophotometer - see instrument section
    3.  pH meter
    4.  Hot plate, 110 V
    5.  Magnetic stirrer
    6.  Pan balance

B,  Reusable Supplies:

    1.  Flasks, volumetric, 50 ml, 100 ml, 1000 ml
    2.  Flasks, Erlenmeyer, 250 ml
    3.  Pipet, volumetric, 25 ml  50 ml
    4,  Pipet, micro, 1  ml graduated 0.1  ml
    5.  Pipet, measuring, 1 ml, 10 ml
    6.  Graduated cylinders, 500  ml, 100 ml,  50 ml,  25 nil
    7.  Beakers, 250 ml
    8.  Funnel, 80 mm diameter
    9.  Ring stand and 3 inch ring
   10,  Watch glass
   11.  Anion and cation exchange resin cartridges
   12.  Reagent bottles
   13.  pH paper
   14.  Specialized glassware - See apparatus section

C.  Consumable Supplies:

    1.  Reagents, analytical  reagent grade

        a.  Arsenic trioxide
        b.  Selenium metal
        c.  Zinc metal  (200 mesh)
        d.  Potassium iodide
        e.  Stannous chloride
        f.  Sulfuric acid 18N
        g.  Hydrochloric  acid,  concentrated
        h.  Nitric acid
        i.  Perchloric acid,  70-72 percent
        j.  Sodium hydroxide

    2.  Gases

        a.  Argon
        b.  Hydrogen
                                                                        F9-3

-------
      WATER  MONITORING  PROCEDURE:  Determination of Arsenic and Selenium
      1.   Analysis  Objective:

          To  determine  the arsenic and  selenium concentration, as listed in the
          Interim  Primary Drinking Water Regulations, in water samples from
          potable water treatment and distribution facilities.

      2.   Brief  Description of  the Analysis:

          Samples are prepared  to distinguish between inorganic and total (inorganic
          and organic)  metal  by  appropriate acid digestion.  Either one of the
          prepared  samples or standards are treated with SnCl^, a reducing agent to

          convert the metal to  its lowest oxidation state, Zinc is added to the acidi-
          fied sample,  generating hydrogen and producing the evolution of the metal
          hydride which is aspirated into a Argon-Hydrogen flame of an atomic absorption
          spectrophotometer.  Calculations are made from a standard curve by measuring
          the peak  heights of the samples and reading the concentration.

      3.   Applicability of  this Procedure:

          a.   Range of  Concentration -  From 2 to 20 yg/1 is  the working  range of the
              method.
          b,   Pretreatment  of Samples  - Upon collection the  pH of the  sample should
              be lowered to below 2 by  the addition of concentrated nitric acid.
              The maximum holding time  is 6 months.
          c.   Treatment of  Interferences  -  In analyzing most surface and ground waters,
              interferences are rarely  encountered.  Organic forms of  arsenic and
              selenium  must be  oxidized before analysis,

          d.   Source of Procedure - Methods for  Chemical Analysis of Water and Wastes,
              1974, Environmental Monitoring and Supply Laboratory, Cincinnati, Ohior.
E9-4

-------
WATER MONITORING PROCEDURE;  Determination of Arsenic and Selenium

Operating Procedures:
A.  Apparatus and Instrumentation
B.  Instrument Calibration
C.  Reagent Preparation
D.  Sample Preparation
E.  Standards Preparation
F.  Samples and Standards Treatment
G.  Calculations
                                                                      E9-5

-------
 WATER  MONITORING PROCEDURE:   Determination  of  Arsenic and Selenium
                                                                                                         E9-6
 OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Apparatus and
   Instrumentation
1.  Atomic absorption  spectro-
   photometer.
                           2.  Hollow  cathode  lamp.
                           3.  Recorder.
                          4.  Flowmeter.
                           5. Medicine dropper.
                           6. Reaction flask.
                          7. Special gas inlet-outlet
                             tube.
                          8. Magnetic stirrer.



                          9. Drying tube.
 la. Any atomic absorption spectrophotometer that
    allows the introduction of a gaseous sample;
    double-beamed instrument preferred because of
    increased stability.

 2a. Arsenic or selenium hollow cathode lamp compati-
    ble with the spectrophotometer.

 3a. Any variable-speed recorder that is compatible
    with the spectrophotometer; a chart speed of
    10 mm/minute recommended.

 4a. A flowmeter capable of measuring 1 I/minute, such
    as a Gilmont No. 12 or equivalent used for
    auxiliary argon.

 5a. A dropper capable of delivering 1.5 ml, fitted
    into a size "0" rubber stopper.

 6a. A pear-shapped vessel with side arm and 100 ml
    capacity, both arms having 14/20 joint
    (Scientific Glass JM-5835)  or equivalent.

 7a. Constructed from a micro cold finger condenser
    (Scientific Galss JM-3325)  or equivalent by
    cutting off the portion below the 14/20 ground
    glass joint.

8a. Must be strong enough to homogenize a 50% mixture
    of zinc dust and water.
8b. See reagent section.

9a. 100 mm long polyethylene tube filled with  glass
    wool  to keep particulate matter out of the
    burner.

-------
WATER MONITORING PROCEDURE:   Determination  of Arsenic  and  Selenium
 OPERATING  PROCEDURES
A. Apparatus and
   Instrumentation
   (Continued)
        STLF SEQUENCE
10. Apparatus setup:   See
    schematic.
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
lOa.  Connect the apparatus  with  the burner of the
     spectrophotometer as indicated in the schematic.
     Connect the outlet of  the reaction vessel  to  the
     auxiliary oxidant input of  the burner with Tygon
     tubing.  Connect the inlet  of the reaction
     vessel  to the outlet side of the  auxiliary
     oxidant, argon supply, control  valve of the
     instrument.
 TRAINING
GUIDE NOTES
B. Instrument
   Calibration
    Prepare atomic absorption
    spectrophotometer for
    operation.
                              Install  the hollow cathode
                              lamp for the element being
                              measured (As or Se)  in  the
                              instrument, set the  wave-
                              length dial and align the
                              lamp in accordance with
                              the manufacturer's
                              instructions.

                              Set the slit width accord-
                              ing to the manufacturer's
                              suggested setting for the
                              element being measured
                              (As or Se}.

                              Turn on the instrument  and
                              ajust the current to the
                              hollow cathode lamp  as
                              suggested by the
                              manufacturer.
 la.  It is not possible to formulate instructions
     applicable to every instrument but in general
     one can follow the step sequence in the
     adjacent column and by consulting the instruction
     manual  for your particular instrument.

 2a.  Arsenic wavelength 193.7 mm.
     Selenium wavelength 196.0 mm.
 2b.  See instruction manual for your particular
     instrument.
                                 4a.  Allow the instrument to  warm up,  usually  10-20
                                     minutes,  until  the  energy  source  stabilizes.
                                                                                                           E9-7

-------
                                                                                                           E9-8
 WATER  MONITORING PROCEDURE:   Determination of Arsenic and Selenium
 OPERATING PROCEDURES
                               STEP SEQUENCE
                                    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                        TRAINING
                                                       GUIDE NOTES
B.
Instrument
Calibration
(Continued)
5. Install  a Boling burner
   head.

6. Turn on the argon and
   adjust to a flow rate of
   about 8 1/minute with
   the auxiliary argon flow
   at 1 1/minute.

7. Turn on the hydrogen,
   adjust the flow rate of
   about 7 1/minute and
   ignite the flame which
   will be essentially color-
   less.

8. Adjust the burner head
   both sideways and verti-
   cally in the 1ight path
   until maximum response is
   obtained by atomizing a
   freshly prepared standard
   solution for the element
   being measured (1.00 ml
   equals 1.00 mg (As or Se))
                                                           7a.  The gas  flow rates  may  require adjustment to
                                                               optimize the flame  for  your particular instrument
                                                           8a.  The instrument is  now ready  to  run  standards  and
                                                               samples.
C.  Reagent Preparation
   1.  Dionized
      Distilled Water
   2.
   Nitric  Acid
   Concentrated
   (HNOJ
1.  Prepare by passing dis-
   tilled water through a
   mixed bed of cation and
   am"on exchange resin.

1.  Commercially available
   reagent grade.
                                                        la.  Use  deionized  distilled water  for  the  preparation
                                                            of all  reagents, calibration standards, and  as
                                                            dilution water.
la. If metal  impurities are found to be present,
    distill  reagent grade nitric acid in a
    borosilicate glass distillation apparatus.

-------
WATER MONITORING PROCEDURE:
 Determination of Arsenic and Selenium
 OPERATING PROCEDURES
    STLP SEQUENCE
                                                           INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                       TRAINING
                                                      GUIDE NOTES
C. Reagent Preparation
   (Continued)
   3. Hydrochloric Acid
   4. Diluent - Stock
5. Potassium Iodide
   Solution
6. Stannous Chloride
   Solution
Commercially avialagle
reagent grade.

Add 100 ml  18 N Sulfuric
acid and 400 ml hydro-
chloric acid to 400 ml
deionized distilled water
in a 1 liter volumetric
flask and bring to volume
with deionized distilled
water.
                           1. Weigh 20 grams potassium
                              iodide, KI, on a pan
                              balance.

                           2. Transfer the reagent into
                              a 250 ml Erlenmeyer flask
                              and dissolve with 100 ml
                              deionized distilled water
                              Weigh 100
                              chloride,
                              balance.
          grams
          SnCl~,
                                           stannous
                                            on  a  pan
                              Transfer the reagent into
                              a 250 ml Erlenmeyer flask
                              and dissolve with 100 ml
                              concentrated hydrochloric
                              acid, HCL.
                                                        la. Use a 100 ml and 500 ml graduate.
                                                        Ib. Diluent used for preparation of working standards
                             la.  Use a weighing disk.
                             2a. Use a 100 ml  graduate.
la.  Use a weighing disk.
                             2a. Use a 100 ml  graduate.
                                                                                                            E9-9

-------
WATER MONITORING PROCEDURE:   Determination  of Arsenic  and  Selenium
                                                                                                       E9-10
 OPERATING PROCEDURES
STEP SEQUENCE
                                   INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reagent Preparation
   (Continued)
   7.  Stock Arsenic
      Intermediate
      Arsenic  Solution
1. Weigh 4 grams sodium
   hydroxide, NaOH, on a pan
   balance.

2. Transfer the reagent to a
   250 ml beaker and add 100
   ml of deionized water,
   allow to dissolve.

3. Weigh 1.3209 grams arsenic
   trioxide, As?0_ on an

   analytical balance.

4. Transfer the reagent to
   the beaker containing the
   NaOH solution, allow to
   dissolve.

5. After dissolution transfer
   into a clean 1000 ml
   volumetric flask and
   dilute to the mark with
   deionized distilled water.

1. Pipet 1  ml stock arsenic
   solution (1000 mg/1) into
   a 100 ml volumetric flask
   and bring to volume with
   deionized distilled water
   containing 1.5 ml con-
   centrated nitric acid per
   liter.
                        la. Use a weighing disk.
                                                           2a.  Stir  if  necessary with a  glass or plastic rod,
                                                           3a.  Using  a weighing  disk.
                                                           4a.  Stir  if  necessary.
                                                           5a. Use a  plastic wash bottle to rinse the beaker
                                                              and stirring rod  into  the volumetric  flask.
                                                           5b. One ml  equals 1.00 mg  As (1000 mg/1).
                        la. One ml equals 10.0 mg As (10 mg/1).
                        Ib, Use a volumetric pi pet.
                        lc. This solution is made up fresh at time of use.

-------
WATER MONITORING PROCEDURE:   Determination of Arsenic and Selenium
 OPERATING  PROCEDURES
       STEP SEQUENCE
                                INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                        TRAINING
                                                       GUIDE NOTES
C. Reagent Preparation
   (Conti nued)
   9. Standard Arsenic
      Solution
  10. Perchloric Acid
      70-725;; HCIO,
   Pi pet 10 ml  intermediate
   arsenic solution in 100
   ml  volumetric  flask and
   bring to volume with
   deionlzed distilled water
   containing 1.5 ml  con-
   centrated nitric acid per
   liter.

   Commercially available
   grade.
                             la.  One ml  equals  1  pg As  (1
                             Ib.  Use a volumetric pipet.
                             Ic.  This solution  is made  up
                             mg/1).

                             fresh at time of use,
D. Sample Preparation
   1. Inorganic Arsenic
      or Selenium
   2, Total (Inorganic
      and Organic)
      Arsenic
1.
To a 50 ml  volumetric
flask add 25 ml  of water
sample, 20 ml  concentrated
                              HC1 and 5 ml
                of 18N H2S04.
1. To 50 ml  of water sample
   in a 150 ml beaker add
   10 ml of concentrated
   nitric  acid, and 12 ml of
   18 N sulfuric acid.  This
   mixture is evaporated to
   SO,, fumes (a volume of

   about 20 ml).

2. To maintain oxidizing
   conditions add small
   amounts to nitric acid
   whenever the red-brown
   NO- fumes disappear.
la. Allow to cool  to ambient temperature.
Ib. Use a 25 ml  volumetric pipet,
Ic. Use 10 ml  measuring pi pets.
                             la.  Oxidizing conditions must be maintained at all
                                 times to avoid loss of arsenic.
                                                           2a. Add nitric acid in
                                                               measuring pipet.
                                                     1  ml  increments using a
                                                                                                                 E9-11

-------
                                                                                                        E9-12
WATER MONITORING PROCEDURE:   Determination of Arsenic and Selenium
 OPERATING  PROCEDURES
       STEP SEQUENCE
   INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
0. Sample Preparation
   (Continued)
3. Allow to cool  slightly,
   add 25 ml  of deiom'zed
   distilled water and 1  ml
   perchloric acid, and
   again evaporate to SO,
   fumes.

4. Allow to cool,  add 40  ml
   concentrated HCL, trans-
   fer the solution into  a
   100 ml volumetric flask
   and dilute to the mark
   with deionized  distilled
   water.
3a. Use a 25 ml  graduate.  Use a 1  ml  measuring
    pipet.
                                                           4a.  Use  a  plastic wash  bottle  to  rinse  the beaker
                                                               during the  transfer.
                                                           4b.  Use  a  50  ml  graduate.
E. Standards
   Preparation
   Transfer 0,  0.5,  1.0,  1.5
   and 2.0 ml  standard
   arsenic or  selenium so-
   lution to 100 ml  volumet-
   ric flasks  and bring to
   volume with  diluent to
   obtain a concentration of
   0,  0,  10, 15,  and  20 pg/1
   arsenic or  selenium.
la. Use a 1  ml  micro pipet graduated in 0.1  ml.
Ib. Refer to reagent preparation section.
F.  Samples and
   Standards  Treatment
   Transfer  a  25  ml  aliquot
   of sample prepared  as  in
   (D.I)  or  0.2)  or  standard
   prepared  as  in (E.)  to a
   reaction  vessel.
la.  Use a 25 ml  volumetric pipet.

-------
WATER MONITORING PROCEDURE:   Determination  of  Arsenic  and  Selenium
OPERATING PROCEDURES
       STEP SEQUENCE
   IMfORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F. Samples and
   Standards Treatment
   (Continued)
2. Add 1  ml  of 20 percent
   potassium iodide to
   arsenic samples and
   standards only.

3. Add 0.5 ml  of 100 percent
   stannous  chloride
   solution.
                           4. Attach the reaction vessel
                              to the special  gas inlet-
                              outlet glassware.

                           5. Fill  the medicine  dropper
                              with  about 1-1/2 ml of
                              zinc  slurry that has been
                              kept  in suspension with a
                              magnetic stirrer.

                           6. Firmly insert the  stopper
                              containing the  medicine
                              dropper into the side
                              neck  of the reaction
                              vessel.

                           7. Squeeze the bulb to intro-
                              duce  the zinc slurry into
                              the sample or standard.

                           8. When  the absorbance
                              reaches its maximum and
                              the recorder pen returns
                              part  way to the base line,
                              remove the reaction vessel
2a. Us a 1  ml  measuring pi pet.
2b. Omit potassium iodide for selenium determinations
                                                           3a.  Use  a  1  ml  measuring  pipet.
                                                           3b.  Allow  at least 10  minutes  for  the  arsenic  or
                                                               selenium to be reduced  to  its  lowest  oxidation
                                                               state.
                                7a. The argon carrier gas will  sweep the generated
                                    metal  hydride into the burner.   The metal  hydride
                                    will produce a peak almost immediately.
                                                                                                              E9-13

-------
                                                                                                         E9-14
WATER MONITORING PROCEDURE:    Determination of Arsenic  and  Selenium
OPERATING PROCEDURES
       STEP SEQUENCE
   INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G. Calculations
1. Draw a standard curve by
   plotting peak heights of
   standards verus concen-
   trations of standards.

2. Using your standard curve
   measure the peak heights
   of the samples  and deter-
   mine the concentration
   from the curve.
la.  Use linear graph paper.
                                                           2a.  Take  in  account any dilution factors.   In  this
                                                               case  the sample was diluted 1 +  1 with  acid,
                                                               multiply the concentration obtained from the
                                                               curve by two.

-------
WATER MONITORING PROCEDURE:   Determination  of  Arsenic and Selenium
ft
Flow
Meier
JM -33215 -.
Medicmc ^y"-xo
Dropnei in K.. —'--
Size "0" V?~
Rubber \^/
Stopper
tryon
1 . Drying p-*^1
1 TU')C t 	 i .. - A-,
J _
-------
                          A  PROTOTYPE  FOR  DEVELOPMENT OF
                          ROUTINE  OPERATIONAL  PROCEDURES

                                     for the

                   DETERMINATION OF NITRATE-NITRITE NITROGEN AND
                   OF  NITRATE NITROGEN,  CADMIUM REDUCTION METHOD
                                   as applied in

                          WASTEWATER TREATMENT FACILITIES
                                    and in the
                        MONITORING OF EFFLUENT WASTEWATERS
                                 Developed by the
               National Training and Operational Technology Center
                        Office  of Water  Program Operations
                        U.S.  Environmental  Protection  Agency
CH.N.n/n.EMP.lb.3.80

                                                                        E10.A-1

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen
                                and of Nitrate Nitrogen, Cadmium Reduction
                                Method
This operational procedure was developed by:
   NAME       Don Roach
   ADDRESS    Miami-Dade Community College, South Campus, 11011  S.W.  104 Street,
              Miami, Florida 33176
   POSITION   Chairman - Chemistry Department
   EDUCATION AND TECHNICAL BACKGROUND
              B.S. - Chemistry
              M.S. - Chemistry
              PhD. - Analytical  Biochemistry
              1  year Commercial  Laboratory Chemist
              10 years College Chemistry Instructor
              7  years Chemical Consultant to Industry
                                                                     E10.A-3

-------
   EFFLUENT  MONITORING  PROCEDURE:   Determination  of  Nitrate-Nitrite  Nitrogen
                                   and  of Nitrate Nitrogen,  Cadmium  Reduction
                                   Method
   1. Objective:

     To  determine  the  nitrate-nitrite  nitrogen  and  the  nitrate  nitrogen  content
     of  an  effluent.

   2. Brief  Description of  Analysis:

     The procedure converts  nitrate  nitrogen  to  nitrite  nitrogen when  the
     nitrate  is  passed through  a  column  containing  copper-cadmium granules.
     Nitrate  is  almost quantitatively  reduced to nitrite by  this process.
     The resulting nitrite is determined by reacting  the effluent with sul-
     fanilamide  and coupling with N  -  (1-napthyl) - ethylenediamine  dihydro-
     chloride to form  a highly  colored dye which can  then be determined
     colorimetrically.   A  correction must be  made for any nitrite initially
     present  in  the sample since  the method determines  total  nitrite.  The
     concentration of  nitrite originally present in a sample can be  determined
     by  omitting the initial copper-cadmium reduction and carrying out the
     remainder of  the  procedure.   Separate nitrate-nitrite values for  a  sample
     may be obtained by analyzing two  aliquots  of the sample; one with the
     copper-cadmium reduction step and one without  the  initial  reduction step.

   3. Applicability of  this  Procedure:

     a.  Range of Concentration:

         0.01  to  1,0 mg N03-NOZ  N/liter

         (The  range may be  extended for samples by dilution.)

     b.  Pretreatment of Samples:

         The Federal  Register Guidelines  do not specify any pretreatment.

     c.  Treatment  of Interferences in  Samples:

         This  procedure includes directions for removal of turbidity and/or of
         grease and oil  from  samples.   It  also includes addition of EDTA to
         eliminate  interferences from metals.  No other Interferences are noted
         in  the Source  of Procedure.*
 * Source  of Procedure:   Methods  for Chemical  Analysis  of Water and  Wastes,  1979,
   U.S.  Environmental  Protection  Agency,  Enivornmental  Flonitoring and Support
   Laboratory,  Cincinnati,  Ohio,  page 353.3-1  (Issued 1974).
nn.A-4

-------
 EFFLUENT MONITORING  PROCEDURE:
Determination of Nitrate-Nitrite Nitrogen
and of Nitrate Nitrogen, Cadmium Reduction
Method
 FLOW SHEET:
                TURBIDITY REMOVAL
                  (if necessary)
                OIL AND GREASE REMOVAL
                    (if necessary)
                ADJUST pH TO BETWEEN 5
                AND 9   (if necessary)
                REDUCTION OF NITRATE TO
                NITRITE BY PASSING THROUGH
                CADMIUM REDUCTION COLUMN
                COLOR DEVELOPMENT BY REACTION
                OF NITRITE WITH SULFANILAMIDE
                AND WITH N-(l- NAPTHYL) -
                ETHYLENEOIAMINE DIHYDROCHLORIDE
                MEASUREMENT OF ABSORBANCE AT
                540 nm
                RESULTS:  NITRATE N PLUS
                ORIGINAL NITRITE N
The above procedures determine nitrate N plus nitrite N.  The initial  nitrite
concentration of the samples could  be determined without reduction.   Thus,
the nitrate concentration can be determined by:

Nitrate N = Total Nitrite N - Nitrite N without reduction
                                                                       E10.A-5

-------
     EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen
                                     and of Nitrate Nitrogen, Cadmium Reduction
                                     Method


     Equipment and Supply Requirements

     A. Capital Equipment:

        1. Balance, analytical, 160 g capacity, precision +^0.1 mg
        2. Balance, triple  beam, 500 g capacity, precision +_ 0.25 g
        3. pH meter/combination electrode, range 0-14 pH
        4. Refrigerator, temperature range 2° - 10°C
        5. Spectrophotometer, wave length range 325-825 nm
        6. Still a,nd de-ionizing cartridges (or other means  of distilling and
           de-ionizing water)

     B. Reusable Supplies:

        1. One apron, laboratory
        2. One 100 ml beaker
        3. Four 250 ml beakers  (3 for buffer solutions)
        4. One 400 ml beaker
        5. One 1 liter beaker
        6. One 2 1iter beaker
        7. Two bottles, Barnes with stoppers and  two  droppers,  small  gauge
        8. One 150 ml bottle, dropper
        9. One 250 ml bottle, plastic wash
        10. One 100 ml bottle, storage with screw-on cap (storage of 6N HC1)
        11. Seven 1 liter bottles, storage, brown with screw-on caps or rubber stoppers
        12. Two 5 gallon bottles, water with bottom spout
        13. One brush, camel hair (cleaning analytical balance)
        14. Two brushes, bottle  (cleaning glassware)
        15. One bulb, propipet type
        16. One buret holder, double clamps (reduction column support)
        17. Two columns, reduction (see Figure 1 at the end of this section)
        18. Three cuvettes
        19. One 25 ml cylinder, graduated
        20. One 50 ml cylinder, graduated
        21. One 100 ml cylinder, graduated
        22. One 500 ml cylinder, graduated
        23. One 1 liter cylinder, graduated
        24. One 50 ml flask, volumetric with stopper (dilution of sample)
        25. Twelve 100 ml flasks, volumetric with stoppers (for standards)
        26. X 100 ml flasks, volumetric with stoppers  (for samples - 1 flask
           per sample)
        27. Twelve 250 ml flasks, Erlenmeyer with stoppers (for standards)
        28. X 250 ml flasks, Erlenmeyer with stoppers  (for samples-1 flask per sample)
        29. One 1 liter flask, Erlenmeyer, or a large, empty  chemical  bottle
           (for Cd washings)
        30. Three 1 liter flasks, volumetric with stoppers
        31. Two 2 liter flasks, volumetric with stoppers
        32. One filter funnel  for 0.45 u filter (turbidity removal)
E10.A-6

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen
                                and of Nitrate Nitrogen, Cadmium Reduction
                                Method

B. Reusable Supplies (Continued)

  33. One funnel, powder
  34. One funnel, large powder with large filter paper (for Cd washings)
  35. One 250ml  funnel, separatory (oil  and grease removal)
  36. One pair glasses, safety
  37. Two  hoses,  rubber, 3" strip, 4 cm 1.0. with screw type clamp
  38. One notebook (recording data)
  39. Two 100 ml  volumetric pipets (construction of reduction columns)
  40. One C,5 ml  pipet, volumetric
  41. One 1 ol pipet, volumetric
  42. One 2 ml pipet, volumetric
  43. One 5 ml pipet, volumetric
  44. One 10 ml  pipet,  volumetric
  45. One 25 ml  pipet,  volumetric
  46. One 50 ml  pipet,  volumetric
  47. One rod, stirring (6" or 12")
  48. One sieve,  40 mesh
  49. One sieve,  60 mesh
  50. One spatula (scoopula )
  51. Two stands, ring  (support funnel, and reduction column)
  52. One support, ring, small (support funnel)

C. Consumable Supplies:

   1. Glasswool,  wad
   2. Membrane filter,  0.45 y
   3. Notebook (recording data)
   4. Pen or pencil  (recording data, marking flasks)
   5. Soap
   6. Sponges (for cleaning)
   7. Tissues, soft (wiping cuvettes and electrodes)
   8. Towels, paper
   9. Twelve weighing boats
  10. 26 g ammonium chloride, NH^Cl

 *11. 100ml ammonium hydroxide, NH^OH

 *12. 150 ml buffer solution, STD pH 4
 *13. 600 ml buffer solution, STD pH 7
 *14. 450 ml buffer solution, STD pH 10
**15. 25 g cadmium granules, 40-60 mesh
  16. 55 ml chloroform, CHCU (Freon or another  non-polar solvent may be used.)

  17. 20 g copper sulfate, pentahydrate, CuSO^-SH^O

  18. 3.4 g disodium ethylenediamine tetraacetate, ^]o^i4^2^a2^8
  19. 1  g N-(l-napthyl) - ethylenediamine dihydrochloride, C12H^N2'2HC1

 *20. 200 ml hydrochloric acid, concentrated, HC1
  21. 100 ml hydrochloric acid, dilute (6N), HC1
  22. 100 ml phosphoric acid, concentrated, H3P04

 *23. Potassium dichromate (cleaning solution),  K^r^Qj

  24. 7.218 g potassium nitrate, KN03
                                                                        E10.A-7

-------
       EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen
                                      and of Nitrate Nitrogen, Cadmium Reduction
                                      Method

       C.  Consumable Supplies  (Continued)

        25. 6.072 g potassium  nitrite,  KNO«

        26, 240  g sodium hydroxide, pellets, NaOH
        27. 10 g sulfanilamide, CgHgN^S

       *28, Sulfuric acid, concentrated,  (cleaning solution) H-SQ^

        29. 100  g zinc  sulfate, heptahydrate, ZnSO.-7HpO

        30. Labels, package,  1 1/2 x  1  inch

        31. Paper,,graph 8 1/2 x 11,  package

       All  reagents should be  reagent  grade.

       The  above  amounts do not allow  for  spillage or mistakes.
       "These amounts will vary

      **MCB  Reagents  .  .  .
E10.A-8

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen
                         and of Nitrate Nitrogen, Cadmium Reduction
                         Method
  10 cm
            80-85  ml
  25 cm
    ,
                         •3 cm I.D.
                      3.5  mm I.D.
                      Cu/Cd
                  100 mi
                volumetric
                   pipet
                     GLASS WOOL PLUG

                            CLAMP
TYGON TUBING
                                                     -Cut
                                                •Cut
             Figure 1. Reduction column
                                                       E10.A-9

-------
 EFFLUENT  MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and of
                            ~~  Nitrate Nitrogen, Cadmium Reduction Method
                                                                                                            E10.A-IO
 OPERATING  PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
DETERMINATION OF NITRATE
A. Equipment
   Preparation

   1.  Glassware Wash-Up
   2.  Balance Inspection

   3.  Spectrophotometer
      Inspection
NITRITE NITROGEN AND OF NITRATE  NITROGEN,  mg/liter
  1,  Clean all  glassware  in
     suitable detergent.
  1,  Clean  balance.

  1.  Clean  spectrophotometer.

  2,  Turn power  on  by  rotating
     the  power control
     clockwise.

  3.  Select wavelength  by
     rotating the wavelength
     control  knob either
     direction until the  proper
     wavelength  is  reached.

  4.  Zero the instrument  by
     bringing the meter needle
     to  "0" on the  percent
     transmittance  scale.

  5,  Use  an empty cell  and
     adjust the light control
     to 100% T.
                                                       I
                                                       (p. 41 )
la. Distilled water drains without leaving any
    droplets on surfaces.
Ib. Use chromerge if necessary.

la. Free of dust and dirt.

la. Free of dust and dirt,

2a. Pilot lamp on.
2b. Directions are for Spectronic 20.


3a. 540 nm on the wavelength scale.
                                                          4a.  Meter needle reads zero.
                                                          5a. To  be  sure  that  the  Instrument  can achieve  100%  T.

-------
 EFFLUENT MONITORING PROCEDURE:
Determination of Nitrate-Nitrite Nitrogen and of
Nitrate Nitrogen, Cadmium Reduction Method
 OPERATING  PROCEDURES
   STEP SEQUENCE
                                   INFQRMATION/OPERATING GOALS/SPECIFICATIONS
                                                                                        TRAINING
                                                                                       GUIDE  NOTES
B. Reagent Preparation

   1, Distilled Water
   2. Concentrated
      Ammonium Chloride
      EDTA Solution
1.  Prepare approximately ten
   (10)  liters of highly pure
   water.
1.  Weigh 26 g of armnonium
   chloride, NH.C1,  in a
   weighing boat and wash
   into 2.0 liter graduated
   beaker.

2,  Weigh 3.4 g of disodium
   ethylenediamine  tetra-
   acetate, C^mM, .NpNa«Oo,
   and wash into the sane
   beaker.

3,  Add enough distilled water
   to bring the total  volume
   to approximately 1800 ml.

4.  Use a pH meter to adjust
   the pH of the solution to
   8.5 by the dropwise addi-
   tion of concentrated
   ammonium hydroxide,
   NH4OH.

5.  After the pH has been ad-
   justed,  transfer the
   solution to a 2 liter
   volumetric flask.
                          la
                                   An ion exchange column in conjunction with a still
                                   provides an adequate source of highly pure water.
                                   This water will be used for all  reagent prepara-
                                   tion and washing of equipment.
                               Ic.  The pH of the water must be between 5.5-7.5.
                                                          Ib
                          la. Distilled water should be used for all phases of
                              solution preparation including water used in
                              washing a solid into a container.
                                                          4a.  Vix the solution thoroughly by stirring, after
                                                              the addition of each drop of NH.OH.
                                                          5a.  Whenever a solution is transferred, the container
                                                              from which the transfer is made should be washed
                                                              and the washings added to the container to which
                                                              the transfer was made.
                                                                                                            E10.A-11

-------
                                                                                                            E10.A-12
EFFLUENT MONITORING PROCEDURF:  Determination of mtrate-Nitrite Nitrogen and of
                               Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
  Reagent Preparation
  (Continued)
  3, Dilute Ammonium
     Chloride EDTA
     Solution
  4, Color Reagent
 6. Dilute to volume with
   distilled water.

 7. Label the bottle in which
   the solution is stored.

 1. Measure 300 ml  of the
   concentrated ammonium
   chloride-EDTA solution
   into a one liter graduated
   cylinder,

 2. Add  distilled mter to
   bring  the  volume to
   500  ml in  the cylinder.

 3. Swirl  to mix the
   solution.

 4. Store in a labeled
   container.
1. Add 800 ml of distilled
   water to a 1 liter flask,

2. Add 100 ml of concentrated
   phosphoric acid, H3PO., to

   the same flask.

3. Mix thoroughly.

4. Weigh 10 g of sulfanilamfde
   (CgHgN202S) in a weighing
   boat.
6a, The solution is stable for several  months.

                 *

7a, Include the name of the solution, your name and
    the date of preparation.
                                                         4a.  This dilute ammonium chloride-EDTA solution is
                                                             stable for several  months.
la. Use a graduated cylinder.
lb. Use a 1  liter volumetric flask.

-------
EFFLUENT MONITORING PROCEDURE:
     Determination  of Nitrate-Nitrite  Nitrogen  and  of
     Nitrate Nitrogen,  Cadmium  Reduction  Method
OPERATING PROCEDURES
       STEP SEQUENCE
                     INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                                 TRAINING
                                                                GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   5. Zinc Sulfate
      Solution
                           6.
                           7.
                           8.
   Use a wash  bottle  and
   funnel  to wash the
   sulfanilamide into the
   1  liter flask containing
   phosphoric  acid solution.

   Weigh 1 g N-(l-napthyl)-
   ethylenediamine dihydro-
   chloride, Marshall's
   Reagent, and  wash  into
   same flask.

   Dilute to volume with
   distilled water.
   Store in a
   container.
labeled
1,  Weigh 100 g of zinc sul-
   fate heptahydrate,
   ZnSO.-7ILO, in a weighing
   boat.
                           2,
   Wash into a 1
   using a wash
   funnel.
   liter
  bottle
flask
and a
        8a.  Container should be dark 1  liter plastic reagent
            bottle,
        8b.  Store at 4°C when not in use.
        8c.  Use at room temperature.
        8d.  The solution is stable for  several  months.
        8e.  A very faint pink color may show up in this
            color reagent.   You may still  use the reagent.
            If a precipitate forms in the  reagent, though,
            discard  it.

        la.  This reagent is used if flocculation is employed
            as an alternative to filtration if the sample
            requires removal of turbidity.
2a. Use a volumetric flask.
                           3.  Add sufficient distilled
                              water to dissolve all  of
                              the solid.
                                                                                                             E10.A-13

-------
                                                                                                             HO.A-14
 EFFLUENT MONITORING  PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and of
                                        Nitrogen, Cadmium Reduction Method
 OPERATING  PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   6.  Sodium Hydroxide
      Solution (6N)
   7.  Ammonium
      Hydroxide
  8. Hydrochloric
     Acid,  (6N)
4. Dilute to volume with
   distilled water.

5. Store in a labeled
   container.

1. Rapidly weigh 240 g q,f
   solid sodium hydroxide,
   NaOH, pellets 1n a 1 liter
   graduated beaker.
2. Add 500 ml distilled
   water to dissolve the
   sodium hydroxide.

3. Dilute to a total volume
   of 1 liter.

4. Store in a glass bottle or
   jug and stopper with a
   rubber stopper.

5. Label  the container.

1. A 100 ml supply should be
   available.

2. Place in a Barnes
   (dropper) bottle.

1. Add 50 ml of distilled
   water  to a 400 ml beaker.
                                                          5a.  This  solution  is  stable  for  at  least  one  year.
la. This reagent is used if flocculation is employed
    as an alternative to filtration if the sample
    requires removal of turbidity.
Ib. Sodium hydroxide picks up moisture from the air
    quite readily.

2a, The water should be added with constant swirling
    to avoid fusing.  CAUTION:  Heat is liberated.
    Place Beaker in a pan of cold water.

3a. The solution should be allowed to cool  to room
    temperature before the dilution is made.

4e. Sodium hydroxide slowly etches glass causing
    glass stoppers to stick.
4b. The solution is stable for at least a year.
la. Drop quantities may be required for pH
    adjustment.
la. A 100 ml graduated cylinder is suitable for
    measuring the volume of the distilled water.

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and of
                                Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   9.  Copper Sulfate
      Solution (2%)
  10, Nitrate Stock
      Solution
2. Slowly add 50 ml  of
   concentrated hydrochloric
   (HC1) acid (12 N) to the
   same beaker.

3. Mix thoroughly.

4, Store in a TOO ml bottle.

5. Label the container.

1. Weigh 20 g of copper
   sulfate pentahydrate,
   CuSO.-SHpO, in a  weighing

   boat.

2. Wash copper sulfate into a
   two 1iter beaker.
                           3. Add 500 nil  distilled  water
                              and swirl  to dissolve the
                              sol id.

                           4. Add 500 ml  distilled  water
                              and swirl  to nix.

                           5, Store in a labeled
                              container.
1. Carefully weigh 7.218 g of
   potassium nitrate, KNO..,
   in weighing boat.
2a. Measure the acid in a 100 ml  graduated cylinder.
                               3a. Use a graduated cylinder to measure 500 ml.
                               5a.  This  solution is  stable for at least one year.
la. An analytical balance should be used.
                                                                                                             E10.A-15

-------
                                                                                                             E10.A-16
 EFFLUENT MONITORING PROCEDURE.:  Determination of Nitrate-Nitrite Nitrogen  and  of
    — 		                    Nitrate Nitrogen, Cadmium Reduction  Method
 OPERATING  PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
  11.  Nitrate Standard
      Solution
2. Transfer the solid to a
   1 liter volumetric flask
   equipped with a powder
   funnel.

3. Use wash bottle to wash
   the solid into the flask.
                           4,  Add sufficient distilled
                              water to dissolve the
                              solid,

                           5.  Dilute to volume with
                              distilled water and
                              thoroughly mix.

                           6.  Store in a labeled glass
                              bottle.

                           7.  Preserve the solution by
                              adding 2 ml of chloroform,
                              CHC13,
1. Carefully pipet 10.0 ml of
   nitrate stock solution
   into a 1  liter volumetric
   flask.
                           2, Dilute to volume with
                              distilled water.
2a. This is best achieved by washing the solid onto
    the funnel with a wash bottle.
3a. The weighing boat should be rinsed three times
    and all of the rinse water should be added to
    the flask.

4a. About 500 ml is sufficient.
7a. The solution prepared^stored and preserved in
    this manner should be stable for at least
    6 months.
7b. The nitrate stock solution contains 1.00 mg of
    nitrate nitrogen (NO-j-N) in each 1.00 ml of

    solution.

la. This nitrate standard solution should be prepared
    fresh for each use.
Ib. The nitrate stock solution should be at room
    temperature before using.
Ic. Use a 10 ml volumetric pipet.

-------
EFFLUENTMONITORING PROCEDURE.;  Determination  of  Nitrate-Nitrite  Nitrogen  and  of
                            ~~   Nitrate  Nitrogen, Cadmium Reduction  Method
OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
  12.  Nitrite Stock
      Solution
3. Store in a labeled
   container.
1. Weigh 6.072 g of
   potassium nitrite, KNCL,

   in a weighing boat.

2. Transfer the solid to a
   1 liter volumetric flask
   using a powder funnel.

3. Use wash bottle to wash
   the solid into the flask.

4. Add sufficient distilled
   water to dissolve the
   solid.

5. Dilute to volume and mix
   thoroughly.

6. Store in a labeled glass
   bottle.

7. Preserve the solution by
   adding 2 ml  of chloroform
   for each 1 liter of solu-
   tion and refrigerate when
   not in use.
3a. Use within two hours of preparation.
3b. The nitrate standard solution contains 0.01 rug
    of nitrate nitrogen (NO_-N) in each 1.0 ml of
    solution.              J

la. An analytical  balance should be used for all
    weighings involving standards.
                                                          3a.  The  weighing  boat should be washed three times
                                                              and  the washings  added to the flask.

                                                          4a.  About 500 ml  is  sufficient.
                                                          7a.  The  solution  should  be  stable  for  at  least
                                                              3  months  when  preserved this way and  stored
                                                              at about  4°C  when  not in  use.
                                                          7b.  The  nitrite stock  solution  contains 1.00 mg  of
                                                              nitrite nitrogen (NO?-N)  in each 1.0  ml  of
                                                              solution.
                                                                                                             E10.A-17

-------
                                                                                                          E10.A-18
 EFFLUENT MONITORING  PROCEDURE:
     Determination of Nitrate-Nitrite Nitrogen and of
     Nitrate Nitrogen, Cadmium Reduction Method
 OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE NOTES
B, Reagent Preparation
   (Continued)

  13. Nitrite Standard
      Solution
1. Pipet 10.0 ml of nitrite
   stock solution into a
   1 liter volumetric flask.
                           2. Dilute to volume with
                              distilled water.

                           3. Store in a labeled
                              container.
la. This nitrite standard solution should be prepared
    fresh for each use.
Ib. The nitrite stock solution should be at room
    temperature before using.
Ic. Use a 10 ml volumetric pipet.
                               3a. Use within two hours of preparation.
                               3b. The nitrite standard solution contains 0,01 mg of
                                   nitrite nitrogen (NOj>~N) in each 1.0 ml of
                                   solution.
C.  Reduction Column
   Preparation

   1.  Preparation of the
      Glass Column
1. Construct a glass column
   by joining a 10 cm length
   of 3 cm ID glass tubing
   with a 25 cm length of
   3.5 mm ID tubing using
   figure 1  as a guide.

2. Loosely plug the delivery
   tip of the column with
   glass wool.
la
                                                          Ib
    Figure 1  is at the end of the Equipment and
    Supply Requirements Section.
    The column shown in Figure 1 was constructed by
    cutting both ends off a 100 ml volumetric pipet
    as indicated.
Ic. Fire polish all cut surfaces.

2a. The plug  must be firm enough to hold the cadmium
    granules  in the column, but not so firmly packed
    as to slow down the later flow of solutions
    through the column.

-------
EFFLUENT MONITORING PROCEDURF: Determination of Nitrate-Nitrite Nitrogen and  of
                               Nitrate Nitrogen,  Cadmium Reduction  Method
OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reduction Column
   Preparation
   (Continued)

   2. Preparation of
      Copperized
      Cadmium for
      Packing the
      Glass Column
1.  Weigh about 25 g of
   cadmium granules in a
   weighing boat.
                           2.  Transfer the  cadmium  to  a
                              400 ml  beaker.

                           3.  Add enough  dilute  (6N)
                              hydrochloric  acid  to
                              cover the granules.

                           4.  Swirl  the contents of the
                              beaker.

                           5.  Pour off the  acid  while
                              retaining the granules
                              1n  the  beaker.   The
                              cadmium should  be  silver.
                           6.  Add  enough  distilled  water
                              to cover  the  granules.
la. This will be enough for one column.
lb. Granulated cadmium (40-60 mesh) can be purchased.
Ic. Alternatively, file sticks of pure cadmium metal
    (reagent grade) with a coarse metal hand file
    (about second cut) and collect the fraction which
    passes a sieve with 10 mesh openings and is re-
    tained on sieves with 40, then 60 mesh openings.
Id. Handling cadmium is hazardous, thus filing should
    be conducted under a hood using rubber gloves and
    mask,

2a. A scupula and wash bottle with water is good for
    this.
                                                                                                                 VIII.C.2.Id
                                                                                                                 (p. 46)
                               5a.  All  decanting  should  be  done  into  a container
                                   equipped  with  a  large funnel  and filter  paper
                                   so as  to  catch all  the small  cadmium  particles.
                               5b.  Use  this  filter  paper for any subsequent cadmium
                                   washings.
                                                                                                               .A-19

-------
                                                                                                           E1Q.A-2Q
EFFLUENT MONITORING PROCEDURE:
       Determination  of  Nitrate-Nitrite  Nitrogen  and of
       Nitrate  Nitrogen,  Cadmium  Reduction Method
OPERATING PROCEDURES
         STEP  SEQUENCE
INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reduction Column
   Preparation
   (Continued)
  7.  Pour off  the water while
     retaining  the granules  in
     the beaker.

  8,  Repeat  steps 6 and 7,
     above,  two more times so
     that the  granules receive
     a total of three dis-
     tilled water washings.

  9.  Add TOO ml of the 2%
     copper sulfate solution to
     the granules and swirl for
     five minutes or until the
     blue color of the copper
     sulfate fades.

10. Carefully decant off the
    solution leaving the
    copperized cadmium
    granules in beaker.

11. Repeat steps 9 and 10
    until  a brown colloidal
    (very fine) precipitate
    of metallic copper does
    form.

12. Wash the copper-cadmium
    at least 10 times  with
    distilled water.

13. Place  the washed copper-
    cadmium on the 60  mesh
    sieve.
                                                         9a. A brown colloidal  (very  fine) precipitate of
                                                             metallic copper  should form.
                                                        lOa. Also decant off through the filter paper any
                                                             precipitate that formed.
                                                        lOb. The cadmium should have a black color.
                                                         lla.  If a brown colloidal precipitate is  formed in
                                                              step 9, and the cadmium is black, do not repeat
                                                              steps 9 and 10.
                                                        12a. All of the brown precipitated copper should be
                                                             removed by washing 10 times but continue to
                                                             wash if any remains.

-------
EFFLUENT MONITORING PROCEDURE: Determination of Nitrate-Nitrite Nitrogen and of
                               Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C.  Reduction Column
   Preparation
   (Continued)
14. Pour water over the
    granules at least three
    times so that al1 the
    small particles will
    wash through the 60 mesh
    screen.

15. Return meshed granules to
    the beaker.

16. Decant off excess water
    used to transfer the
    cadmium.

17, Close the clamp on the
    column delivery tube.

18. Fill  the column with liq-
    uid,  using about 60 ml
    DILUTE ammonium chloride-
    EDTA solution.

19. Loosely fill  the reduction
    column with copper cadmium
    granules to a level  about
    2 cm below the  broad,
    cup-like section as shown
    in Figure 1 page 9.
14a.  Hold the sieve over the filter paper during
     these washings.
                                                         15a.  Use a scupula  and  the wash bottle.
                                                         18a.  Use  a  graduated  cylinder  and  very  slowly  pour  the
                                                              solution  down  the  inside  wall  of this  column so
                                                              air  pockets  do not form.
                                                         19a,  Avoid  tight  packing  of granules  by  allowing  the
                                                              granules  to  "float"  down  through the  solution
                                                              of  ammonium  chloride-EDTA.
                                                         19b.  A glass stirring  rod may  be  used to transfer the
                                                              cadmium to the  column.
                                                                                                             E10.A-21

-------
EFFLUENT HUNITORING PROCEDURE.:  Determination of Nitrate-Nitrite Nitrogen and of
                                Nitrate Nitrogen, Cadmium Reduction Method
                                                                                                            E10.A-22
OPERATING PROCEDURES
         STEP  SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reduction Column
   Preparation
   (Continued)
20. Open the screw clamp and
    measure the flow rate of
    ammonium chloride-EDTA
    solution through the
    column.  The flow rate
    must be between 7 ml  and
    10 ml/minute before you go
    to the next step.  Keep a
    record if you add more di-
    lute ammonium chloride-
    EDTA solution.
                         21. Rinse the column with up ta
                             140 ml dilute ammonium
                             chloride-EDTA solution,
                             draining until  the solutior
                             is about 2.5 cm above_ the
                             top of the granules."  Then
                             close the screw clamp.
20a. To calculate the flow rate, place a 50 or 100 ml
     graduated cylinder under column and measure the
     amount of fluid collected in one minute
20b. The flow rate should be between 7 ml and 10 ml/
     minute.  DO NOT let the column go dry.
20c. If the flow rate is too fast, tighten the screw
     clamp.  If the clamp must be so tight that contra
     is lost, add more copper-cadmium granules to the
     column.
20d. If the flow rate is too slow, decrease the length
     of the copper-cadmium column until a flow rate of
     7-10 ml/minute is achieved.

21a. There is to be a 200 ml rinse with this solution.
     You used about 60 ml in Step 18 and may have
     added more in Step 20.   Now add the balance
     to total 200 ml.
21b. It is convenient to add a second clamp to shut
     off the flow so the flow-regulating clamp can re-
     main undisturbed.
21c. When the column is not  in use, the granules
     should be covered with  solution so they do not
     dry out.
D.  Removal  of
   Interferences

   1.  Turbidity Removal
      (If necessary)
 1.  Prior  to  analysis,  remove
    turbidity from  samples  by
    filtering through a 0.45 u
    membrane  filter.
                                                                                        VI.D
                                                                                        (p.  42)
 la.  If the turbidity is not removed by filtration,
     proceed as follows:  Add 1  ml  of the zinc sulfate
     solution to 100 ml  of sample.   Add enough 6 N
     sodium hydroxide to bring the  pH to 10.5  (about
     8 to 10 drops is usually sufficient).   Let the
     treated sample stand for 15 minutes.  Filter
     through a 0.45 y membrane filter.
 Ib.  Suspended solids can clog the  reduction column.

-------
EFFLUENT MONITORING PROCEDURE:
     Determination of Nitrate-Nitrite Nitrogen and of
     Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Removal  of
   Interferences
   (Continued)

   2.  Oil  and Grease
      Removal  (If
      Necessary)
1. Prior to analysis,
   measure 100 ml of the
   sample (filtered sample
   if the original sample
   was turbid) into a 400 ml
   beaker.

2. By dropwise addition, add
   sufficient concentrated
   hydrochloric acid (12 N)
   to bring the pH down to 2,

3, Place the sample in a
   250 ml separatory funnel.

4. Add 25 ml  of chloroform.

5. Shake gently to extract
   the oils and grease into
   the chloroform layer.
                           6. Allow  the  separatory
                             funnel  to  stand  until all
                             of  the  chloroform  layer
                             settles  to  the bottom.

                           7. Open the stopcock  and
                             allow  the  bottom (chloro-
                             form)  layer to pass into
                             a 400 ml beaker.
la. Oil and grease can clog the reduction column
    and coat the Cu/Cd granules.
                                                          2a.  Use  a  pH meter  in  adjusting  the  pH  to  2,
                                                          2b.  Standardize  using  standard buffer of pH  =  4.00.
4a. Freon or another non-polar solvent may be used.

5a. Carefully release the pressure after shaking
    gently so that no sample is lost.  This can be
    accomplished by inverting the separatory funnel
    and slowly opening the stopcock away from face
    and other people.

6a. Place funnel in ring stand.
6b. Remove stopper while layer is settling.
                               7a.  Grease and  oils  are extracted  into chloroform
                                   layer leaving  a  grease-oil  free  sample  which  is
                                   used  for analysis.
                                                                                                            E1Q.A-23

-------
                                                                                                           E10.A-24
EFFLUENT MONITORING PROCEDURE;  Determination of Nitrate-Nitrite Nitrogen and of
                               Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
D. Removal of
Interferences
(Continued)
E. Preparation of
Nitrate Working
Standards
1. Nitrate Working
Standards

*

STEP SEQUENCE
8. Repeat steps 4, 5, 6, and
7 with 25 ml of fresh
chloroform.

1. Prepare nitrate working
standards by respectively
pipetting the following
volumes of nitrate
standard solution into
each of six 100 ml
volumetric flasks.
Add This For This
Volume of Concentra-
Nitrate tion of
To Flask Standard N03-N in
No. Solution mg/1
1 0.0 ml 0,00
2 0.5 ml 0.05
3 1.0 ml 0.10
4 2.0 ml 0.20
5 5.0 ml 0.50
6 10.0 ml 1.00
2. Dilute each of the flasks
to volume with distilled
water.
3. Stopper and mix thoroughly
by inversion.
INFORMATION/OPERATING GOALS/SPECIFICATIONS
8a. The second chloroform extract is added to the
same beaker as the first extract.

la. Label flasks.
Ib. Use appropriate volumetric pipets (0.5 ml, 1.0 ml,
2.0 ml, 5.0 ml , 10.0 ml).
Ic. The 0.00 solution which contains no nitrate (or
nitrite) serves as the reagent blank for the
nitrate samples and standards which are passed
through the reduction column.



TRAINING
GUIDE NOTES







-------
EFFLUENTMONITORING PROCEDURE
    Determination  of Nitrate-Nitrite  Nitrogen  and  of
    Nitrate  Nitrogen,  Cadmium Reduction  Method
OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E,  Preparation of
   Nitrate Working
   Standards (Continued)
4.  Use the working  standards
   inrmediately after their
   preparation.
F. Reduction of Nitrate
   to Nitrite

   1. Adjustment of pH
   2. Activation of
      Column
   Use a pH meter to adjust
   the pH of each of the
   working standards to
   between 5 and 9 either
   with concentrated hydro-
   chloric acid or with
   concentrated armonium
   hydroxide.
1.  Pipet 25.0 ml  of working
   standard 16 to a small
   Erlenmeyer flask.
                           2. Add 75 ml  of the CON-
                              CENTRATED  ammonium chlo-
                              ride-EDTA  solution to
                              the sane flask.

                           3. Mix the working standard
                              thoroughly by swirling
                              the contents of the flask

                           4. Place a 250 ml beaker
                              under the reduction
                              column.
la. Use a beaker small enough for this volume of
    standard to cover the pH electrode(s).
Ib. **ake sure that the pH meter is calibrated within
    this range.
Ic. Use buffer solutions pH 4, pH 7, pH 10 to
    calibrate and check the meter.
Id. This pH adjustment is necessary to insure that
    the pH is approximately 8.5
    (No pH adjustment is necessary if the pH is
    already between 5 and 9.)

la. Activation of column is necessary to prepare
    surfaces of Cu-Cd granules for reduction process,
Ib. This standard is 1.00 mg NO,-N/liter
    concentration.
Ic. A 25C ml flask is good for this purpose,

2a. A 100 ml graduated cylinder is good for this
    purpose.
                                                                                                            E10.A-25

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and of
                               Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F. Reduction of Nitrate
   to Nitrite
   (Continued)
5. Check that the level of
   ammonium chloride-EOTA
   solution in the column is
   near to the top of the
   granules.

6, Pour the prepared nitrate
   working standard into the
   reduction column.

7. Using the screw clamp
   (see Figure 1) adjust the
   collection rate to
   7-10 ml per minute.
                          8.  Collect the  reduced
                              working standard until
                              the  level of  solution  is
                              one  cm above  the top of
                              the  granules.

                          9.  Close the screw clamp  to
                              stop the flow.

                         10.  Discard the entire re-
                              duced working standard.

                         11-  Measure about 40 ml of
                              DILUTE ammonium chloride-
                              EDTA solution,

                         12.  Pour the 40 ml into the
                              column.
5a. If the level is too high, drain the excess into
    the beaker.
                                                          6a.  Since  the  column  will  not  hold  the  total  amount,
                                                              add  the solution  in portions.
                                                          7a.  The  clamp  should  be  slowly opened  until  a
                                                              collection rate of 7-10 ml  per minute  is
                                                              achieved.
                                                          7b.  A collection  rate of 7-10  ml of  solution per
                                                              minute  should be carefully maintained  throughout
                                                              the  collection process  to  assure complete  re-
                                                              duction of  nitrate in the  sample.
                              lOa.  The  column  is  now  activated.

-------
EFFLUENT MONITORING PROCEDURE:
       Determination  of Nitrate-Nitrite  Nitrogen  and  of
       Nitrate Nitrogen,  Cadmium  Reduction  Method
OPERATING PROCEDURES
         STEP  SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
I  TRAINING
  3UIDE NOTES
F. Reduction of Nitrate
   to Nitrite
   (Continued)
   3. Reduction of
      Working Standards
13.  Collect the solution until
    the level  of the solution
    is one cm above the top of
    the granules.   Then close
    the screw clamp to stop
    the flow.
14.  The column should be ready
    to use.

 1.  Pipet 25.0 ml  of the lowest
    concentration  of nitrate
    working  standard into a
    small Erlenmeyer flask.

 2.  Add 75 ml  of the CONCEN-
    TRATED ammonium chloride-
    EDTA solution  to the same
    flask.

 3.  Mix nitrate working stand-
    ard thoroughly by swirling
    the contents of the flask.

 4.  Place a  short  graduated
    cylinder under the reduc-
    tion column.

 5.  Pour the prepared nitrate
    working  standard into the
    reduction  column.

 6.  Using the  screw clamp (see
    Figure 1)  adjust the col-
    lection  rate to 7-10 ml
    per minute.
13a.  You can check for "carry over"  by collecting
     about 5 to 10 ml  of the final  effluent in a
     clean receiver and adding a  few drops  of the
     nitrite color reagent to verify that no color
     develops after 10 minutes.   (A  very faint pink
     color is negligible).  Repeat  steps 11  through
     13 if significant color develops in this check
     procedure.
 la.  A 250 ml  flask is good for this purpose,
 Ib.  Label the flask.
 Ic.  Begin with the 0.00 mg/liter solution.
                                                          2a.  Use a  100 ml  graduated cylinder,
                                                          4a.  You need to measure 25 ml  of solution  in  the
                                                              graduate.
                                                          5a.  Since the column will  not hold the total  amount,
                                                              add the solution in portions.
                                                          6a.  The clamp should be slowly opened until  a collec-
                                                              tion rate of 7-10 ml  per minute is achieved.
                                                          6b.  A collection rate of 7-10 ml  of solution per  min-
                                                              ute should be carefully maintained throughout the
                                                              collection process to assure  complete reduction of
                                                              the nitrate in the nitrate wor king sTandard.
                                                                                                           E10.A-27

-------
                                                                                                           E10.A-23
 EFFLUENT  MONITORING  PROCEDURE: Determination of Nitrate-Nitrite Nitrogen and of
                        —^—         N-jtr0gen> cadmium Reduction Method
 OPERATING  PROCEDURES
        STEP SEQUENCE
                  INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE NOTES
 F.  Reduction of Nitrate
    to Nitrite
    (Continued)
 7, Discard the first 25 ml
    of solution which is
    collected,

 8. Replace the graduate with
    the rinsed,air-dried
    flask used for this
    standard,

 9. Collect the remaining
    portion of the reduced
    standard in the original
    flask,

10, Analyze the reduced
    standard IMMEDIATELY
    after collection from
    the reduction  column.
                          11. Repeat steps 1  through 10
                              for each of the prepared
                              working nitrate standards,
              7a. This discard portion serves to "wash off"
                  solution remaining in the column from any
                  previous pass-through.

              8a. The solution originally in the flask should now
                  be in the column so you can thoroughly rinse it.
                  A different flask may also be used.
                                                          9a. Close the screw clamp when the level of solution
                                                              is about one cm above the granules.
                                                          9b. About 70 ml should be in the flask.
             lOa. While one solution is passing through the column
                  you should proceed to color development of the
                  previous solution that has already been reduced.
                  Color development (Section G) MUST BEGIN WITHIN
                  15 MINUTES after reduction.

             lla. Proceed from the least concentrated to the most
                  concentrated standard.
             lib. Label each receiver flask.
G. Color Development of
   Reduced Nitrate
   Working Standards
 1.  Use  a  50.0  ml
    remove a  50.0
    from flask  #1
    liter  N03-N).
pi pet to
ml aliquot
(0.00 mg/
la.  By using a propipet the aliquot can remain in
    the pipet during the next two steps.
Ib.  Aliquots of each of the working standards should
    have been passed through the reduction column as
    described in the previous section (Section F).
Ic.  The reduced working standards should be analyzed
    as soon as possible after the reduction and in.
    no case, should they be allowed to stand for more
    than 15 minutes after reduction before color
    development is begun.

-------
EFFLUENT MONITORING PROCEDURE:
    Determination of Nitrate-Nitrite Nitrogen and of
    Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP  SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G.  Color Development of
    Reduced Nitrate
    Working Standards
    (Continued)
2. Discard the remainder of
   the nitrate reduced
   working standard.

3, Shake flask dry.

4. Add the 50.0 ml working
   standard back to same
   flask from which it was
   removed.

5. Add 2.0 ml of the color
   reagent to the 50.0 ml
   of working standard.

6. Mix thoroughly by
   swirling.

7. Allow the  working standard
   to stand  until  color
   develops.
                           8. Repeat steps 1 through 7
                              for each of the reduced
                              working standards.
                                                          3a. Do not rinse the flask.

                                                          4a. If you find the technique in steps 1-4 too
                                                              difficult, transfer the 50.0 ml to a different
                                                              flask.
                                                          5a. Use a 2.0 ml volumetric pipet.
7a. The reduced working standard should be allowed
    to stand for at least 10 minutes but NOT MORE
    THAN TWO HOURS before doing Procedure L,
    Spectrophotometric Measurements.

8a. Start with least concentrated solution and
    proceed to most concentrated.
8b. Rinse the 50.0 ml pipet thoroughly after each
    standard.
H.  Analysis of Samples
   for Nitrate Reduced
   to Nitrite

   1. Dilution of
      Samples (if
      necessary)
   Pipet 25.0 ml  of  unknown
   sample into 50 ml  volu-
   metric flask.
la. Potable water samples will usually require no
    dilution, while sewage samples may require
    dilution.
                                                                                                           no.A-29

-------
 EFFLUENT MONITORING PROCEDURE:  Determination  of  Nitrate-Nitrite Nitrogen and of
                                Nitrate  Nitrogen,  Cadmium Reduction Method
                                                                                                           E10.A-30
 OPERATING  PROCEDURES
                               STEP SEQUENCE
                                   INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE NOTES
 H
Analysis of Samples
for Nitrate Reduced
to Nitrite
(Continued)

2. Adjustment of pH
   3. Reduction of
      Nitrate to Nitrite
      in Samples
   4,  Color Development
      in Samples
2.  Dilute to volume with
   distilled water.
                              Use a pH meter to adjust
                              the pH of each sample to
                              between 5 and 9 either
                              with concentrated hy-
                              drochloric acid or with
                              concentrated ammonium
                              hydroxide.
                        1.  Aliquots  of  each  of  the
                           samples  should  be passed
                           through  the  reduction
                           column as  described  in
                           Procedure  F.3,  "Reduction
                           of Working Standards."

                        1.  Follow the steps  fn
                           Procedure  G,  "Color
                           Development."
                                                          2a. If you need to dilute a sample, you must apply a
                                                              dilution factor to the concentration found from
                                                              a standard curve.
                               la.  Put the  50 ml  of  sample  in  a  small  beaker  so
                                   the pH electrode(s)  is covered  with solution.
                               Ib.  Hake sure  that pH meter  is  calibrated  within
                                   this range.
                               Ic.  Use buffer solutions  pH  4,  pH 7,  pH 10 to  cali-
                                   brate  and  check the meter.
                               Id.  This pH  adjustment is necessary to  insure  that
                                   the pH is  approximately  8,5
                                   (No pH f.rljustpent is  necessary  if the  pH is
                                   already  between 5 and 9.)
                                                   VII.H.I.2a
                                                   (P. 44 )
I.  Preparation of
   Nitrite Working
   Standards

   1,  Nitrite  Working
      Standards
                          Prepare nitrite working
                          standards by respectively
                          pipetting the following
                          volumes of nitrite stand-
                          ard solution into each of
                          six 100 ml volumetric
                          flasks.
                               la,
                               Ib,

                               Ic.
Label flasks.
Use appropriate volumetric pipets (0.5 ml, 1.0 ml,
2.0 ml, 5.0 ml, 10.0 ml).
The 0.00 solution which contains no nitrite (or
nitrate) serves as the reagent blank for the
nitrite standards and samples that are jipt passed
through the column.

-------
EFFLUENT MONITORING PROCEDURE:
    Determination of Nitrate-Nitrite Nitrogen and of
    Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP  SEQUENCE
                 INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE NOTES
 I. Preparation of
   Nitrite Working
   Standards
   (Continued)
   2,  Adjustment of pH
                         o Flask
                           No.
                            1
                            2
                            3
                            4
                            5
                            6
       Add This
       Volume of
       Nitrite
       Standard
       Solution

        0.0 ml
          5 ml
          0 ml
          0 ml
          0 ml
       10.0 ml
For This
Concentra-
tion of
NOz-N in
mg/1

  0.00
  0.05
  0.10
  0.20
  0.50
  1.00
2. Dilute each of the flasks
   to volume with distilled
   water.

3. Use the working standards
   immediately after their
   preparation.

1. Use a pH meter to adjust
   the pH of each of the
   working standards to
   between 5 and 9 either
   with concentrated hydro-
   chloric acid  or with
   concentrated  ammonium
   hydroxide.
             la,
                                                          Ib,
                                                          Ic.
                                                          Id.
Use a beaker small enough for this volume of
standard to cover the pH electrode(s).
Make sure that pH meter is calibrated within this
range.
Use buffer solutions pH 4, pH 7, pH 10 to
calibrate and check the meter.
This pH adjustment is necessary to insure that
the pH is approximately 8.5
(No pH adjustment is necessary if the pH is
already between 5 and 9.)
                                                                                                           E10.A-31

-------
                                                                                                          E10.A-32
EFFLUENT MONITORING PROCEDURE:
      Determination of Nitrate-Nitrite Nitrogen and of
      Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
J. Color Development of
   Nitrite Working
   Standards
 1.  Pipet 25.0 ml  of  each  of
    the nitrite working
    standards  into each  of
    six clean  250  ml
    Erlenmeyer flasks,

 2.  Add 75 ml  of CONCENTRATED
    ammonium chloride-EDTA
    solution to each  of  the
    nitrite working standards,

 3.  Mix each thoroughly  by
    swirling each  flask.

 4.  Use a 50.0 ml  pi pet  to
    remove a 50.0  ml  aliquot
    from  flask #1  {0.00  mg/
    liter N02-N),

 5.  Discard the  remainder  of
    the  standard from the
    flask.

 6.  Shake the  flask dry,

 7,  Add  the 50.0 ml   nitrite
    working standard back to
    the  same flask from which
    it was removed.

8. Add 2.0 ml  of  the color
   reagent to each nitrite
   working standard.

9. Mix thoroughly by
   swirlfng.
la.  Use a 25,0ml  volumetric pipet.
Ib.  Label each flask.
Ic.  The nitrite working standards are not passed
    through the reduction column.
                                                          2a.  Use a  100 ml  graduated cylinder.
                                                          4a.  By  using  a  propipet the  aliquot can  remain in the
                                                              pipet  during  the  next  two  steps.
                                                         6a. Do  not rinse  the  flask.

                                                         7a. If  you find the techniques  in  steps  4-7  too
                                                             difficult, transfer  the  50.0 ml  to a different
                                                             flask.
                                                         8a. Use a 2.0 ml volumetric pipet.

-------
EFFLUENT MONITORING PROCEDURE:
      Determination of Nitrate-Nitrite Nitrogen and of
      Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                    TRAINING
                                                   GUIDE  NOTES
J, Color Development of
   Nitrite Working
   Standards (Continued)
10.  Allow the working stand-
    ards to stand until  color
    develops.

11.  Repeat steps 4 through  10
    for each of the nitrite
    standards.
lOa. At least 10 minutes but NO  MORE  THAN  2  HOURS
     should be allowed before doing Procedure L,
     Spectrophotometric Measurements.

lla. Proceed from the least concentrated to  the most
     concentrated standard.
lib. Rinse the 50.0 ml pipet thoroughly after each
     standard.
K. Analysis of
   Non-reduced
   Samples for
   Nitrite

   1. Dilution of
      Samples (if
      necessary)
   2. Adjustment of pH
   3.  Color Development
 1.  Pipet  25.0 ml  of unknown
    sample into 50 ml  volu-
    metric flask.

 2.  Dilute to  volume with
    distilled  water.

 1.  Use  a  pH meter to  adjust
    the  pH of  each sample  to
    between 5  and  9 either
    with concentrated  hydro-
    chloric acid or with
    concentrated ammonium
    hydroxide.
 1.  Pipet  25.0 ml of sample
    into a clean 250 ml
    Erlenmeyer flask.
 la.  NOTE:   Potable water samples  will  usually
     require no dilution, while  sewage  samples
     may require dilution.

 2a.  If you  need to dilute  a  sample,  you  must  apply  a
     dilution factor to  get a  final answer.
 la,
                                                          Ib,
                                                          Ic.
                                                          Id,
                                                                                                                 VII.K.l.Za
                                                                                                                 (P.  44)
Put the 50 ml of sample in a small beaker so the
pH electrode(s) is covered with solution.
Make sure that pH meter is calibrated within
this range.
Use buffer solutions pH 4, pH 7, pH 10 to
calibrate and check the meter.
This pH adjustment is necessary to insure that
the pH is approximately 8.5.   (No pH adjustment
is necessary if the pH is  already between 5 and 9)
 la.  Use  a  25.0  ml  volumetric  pipet.
 Ib.  Label  the flask.
 Ic.  The  sample  is  not  passed  through  the  reduction
     column.
                                                                                                           E10.A-33

-------
                                                                                                           E10.A-34
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and of
                               Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
K. Analysis of
   Non-reduced
   Samples for
   Nitrite (Continued)
 2. Add 75 ml of the con-
    centrated ammonium
    chloride-EDTA solution to
    the same flask.

 3. Mix the sample thoroughly
    by swirling.

 4. Use a 50.0 ml pipet to
    remove a 50.0 ml aliquot
    from flask,

 5, Discard the remainder of
    the solution from the
    flask.

 6. Shake flask dry.

 7. Add the 50.0 ml  of sample
    back to same flask from
    which it was removed.

 B. Add 2.0 ml  of the color
    reagent to the same flask

 9. Mix the sample thoroughly
    by swi rli ng.

10. Allow the sample to stand
    until  color develops.
                          11.  Repeat  steps  1  through
                              10  for  each  sample.
 Ha. Use a 100 ml graduated cylinder.
                                                          4a.  By using a propipet the aliquot can remain in
                                                              the pipet during the next two steps.
                                                          6a.  Do not rinse the flask.
                                                          8a.  Use a 2.0 ml  volumetric pipet.
lOa.  At least 10 minutes but NO MORE  THAN  2 HOURS
     should be allowed before doing Procedure L,
     Spectrophotometric Measurements.

lla.  Rinse the 50.0 ml pipet thoroughly after each
     sample.

-------
EFFLUENT MONITORING PROCEDURE:
      Determination of Nitrate-Nitrite Nitrogen and of
      Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING COALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
L.  SpectrophotometHc
   Measurements

   1.  Adjusting the
      Instrument
   2.  Reduced Nitrate
      Standards  and
      Sample(s)
      3.  Non-reduced
         Nitrite  Stand-
         ards  and
         Sample(s)
1, Consult the manufacturer's
   instructions for cali-
   brating your particular
   instrument.

2. Adjust the wavelength to
   540 nm.

3. Check to make sure that
   the instrument reads
   infinite absorbance with
   no sample cell in the
   instrument.
1. Use the reduced nitrate
   reagent blank to adjust
   the instrument to zero
   absorbance.

2. Measure and  record the
   absorbance of each re-
   duced nitrate working
   standard.

3. Measure and  record the
   absorbance for each
   reduced sample.

1. Use the nitrite reagent
   blank (non-reduced) to
   adjust the instrument to
   zero absorbance.
la. Instrument must be warmed up for at least
    10 minutes.
Ib. There is an EMP on "Use of a Spectrophotometer."
3a. If it does not read infinite absorbance with no
    sample cell in it, adjust the instrument so that
    it does read infinite absorbance (see manu-
    facturer's instructions),
3b. Use calibration knob to calibrate infinite
    absorbance.

la. Use 0.00 nitrate working standard reagent blank
    which has been passed through the column.
Ib. Adjust to zero absorbance using the calibration
    knob.

2a. Use the nitrate working standards which have
    been passed through the column.
2b. Use data sheet provided.
                                                          3a. Use data  sheet provided.
la.  Use 0.00 nitrite working standard reagent blank.
Ib.  Adjust to zero absorbance using the calibration
    knob.
                                                                                                                 IX.L.2.2b
                                                                                                                 (p.  47)
                                                                                                             E10.A-35

-------
                                                                                                          E10.A-36
EFFLUENT MONITORING PROCEDURE:
      Determination of Nitrate-Nitrite Nitrogen and of
      Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
1. Spectrophotometrlc
   Measurements
   (Continued)
2. Measure and record the
   absorbance of each non-
   reduced nitrite working
   standard.

3. Measure and record the
   absorbance for each non-
   reduced sample.
2a. Use data sheet provided.
                                                          3a.  Use data sheet provided.
IX.L.3.2a
(p.  47)
M. Preparation of
   Calibration Curve
1. Obtain an 8 1/2 x 11 inch
   piece of graph paper.

2, Label the longer side as
   the concentration axis.

3. Label the shorter side
   as the absorbance axis.

4. Use the absorbance value
   and its corresponding ni-
   trate concentration for
   each of the nitrate working
   standards to make a plot of
   absorbance versus
   concentration.

5. On another piece of graph
   paper follow steps 1, 2, 3,
   and 4 using absorbance val-
   ues and the corresponding
   nitrite concentrations for
   each of the nitrite working
   standards.
                                                          2a.  See Training Guide for an example of labeling
                                                              the axis on a calibration curve.
                                                          4a.  Use the  absorbances  and  concentrations  recorded
                                                              on  the data  sheet in Column  B,  "Total  N02+N03-N."

                                                          4b.  This will  be the standard  curve for reduced
                                                              samples.
                                                          5a.  Use the absorbances  and concentrations  recorded
                                                              on  the data  sheet  in Column  D,  "NCL-N."
                                                          5b.  This will  be the standard curve for non-reduced
                                                              samples.
                                                       VII.M.2a
                                                       (p.  45)
                                                       IX
                                                       (p.  <8)
                                                       IX
                                                       (p. 48)

-------
EFFLUENT MONITORING PROCEDURE.:  Determination of Nitrate-Nitrite Nitrogen and of
                               Nitrate Nitrogen, Cadmium Reduction Method
OPERATING PROCEDURES
       STEP  SEQUENCE
                                   INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
N. Checking Column
   Efficiency
1.  Divide the absorbance
   value for the 1.00 rug/
   liter NITRATE (N03)
   working standard by the
   absorbance for the 1.00
   mg/liter NITRITE (N02)
   working standard to
   obtain the column
   efficiency as follows:
                           abs of 1.00 Big/liter N03std x 100 = % efficiency
                           abs of 1.00 mg/liter N02std
                               la.  The abbreviation,  abs is used to stand for
                                   absorbance.
2. Divide the absorbance
   values for each of the
   other NITRATE (N03)

   working standards by the
   absorbance value for the
   corresponding NITRITE
   (NO*) working standard to

   obtain a column efficiency
   value in each case as was
   done in the previous step.

3. Calculate the average
   value for the column
   efficiency.
                                                          2a.
                                   At least one reduced nitrate standard should be
                                   compared to  a nitrite standard of the same con-
                                   centration to check column efficiency, calculated
                                   as given in  Step 1.  If series of the standards
                                   are run, you can calculate the average column
                                   efficiency using this Step 2 and then Step 3.
                                                          3a.
                                                          3b.
                                   The average value for the column efficiency
                                   should be between 96% and 1041.   If the average
                                   % efficiency does not fall in this range,
                                   another cadmium reduction column should be
                                   prepared and tested until the average column
                                   efficiency does fall in this range.
                                   For regeneration of a column, see Training
                                   Guide,
                                                                                      VII.N.3b
                                                                                      (p. 43)
                                                                                                             E10.A-37

-------
                                                                                                            E10.A-38
 EFFLUENT  MONITORING  PROCEDURE:
      Determination of  Nitrate-Nitrite  Nitrogen and of
      Nitrate  Nitrogen, Cadmium  Reduction Method
 OPERATING  PROCEDURES
       STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE NOTES
 0.  Determination  of
    mg/liter  Nitrite
    Nitrogen  Plus  Nitrate
    Nitrogen  in a  Sample
   Use the absorbance for
   the reduced sample and
   the standard curve for
   reduced samples  ("Total
   N02+N03-N") to obtain

   the mg/liter of  nitrite-N
   plus nitrate-N in the
   sample and record it in
   Column (A) on the data
   sheet provided.
 la.
 Ib,
If the sample was not diluted (25 ml of sample
is used), the ing/liter result is read directly
from the nitrate standard curve.
If the concentration of nitrate in the sample
is too high for analysis, the sample must be
diluted.  The procedure is described in H.I and
involves diluting the sample to a 50 ml volume.
In this case, the mg/liter result from the nitrate
standard curve must be multiplied by a dilution
factor which would be:
                                                       IX.O.la
                                                       (p. 47)
                                                              Dilution Factor =
                                                                50ml
                                                     ml sample used in dilution
                                                          Ic,
                                                          Id.
                                   The reduction process converts the nitrate-N
                                   initially present in the sample to nitrite
                                   nitrogen and the species analyzed is nitrite
                                   nitrogen.
                                   Any nitrite nitrogen initially present in the
                                   sample remains as nitrite nitrogen after the
                                   reduction.   Thus the total  nitrite analyzed is
                                   the sum of  the nitrite initially present and
                                   the nitrite which has been  formed by reduction
                                   of nitrate.
                                                       VH.O.lb
                                                       (p. 44)
P.  Determination of
   mg/liter Nitrite
   Nitrogen in a Sample
1.  Use the absorbance for
   the non-reduced sample
   and the standard curve
   for non-reduced samples
   ("N02-N")  to obtain the

   mg/liter of  nitrite-N in
   the sample and  record it
   in  Column  (C) on the data
   sheet  provided.
le. If the sample was not diluted (25 ml  of sample is
    used), the mg/liter result is read directly from
    the nitrite standard curve.
Ib. If the sample was diluted to a 50 ml  volume (as
    given in K.I), the mg/liter  result read from the
    nitrite standard curve must  be multiplied by a
    dilution factor which would  be:
                                                   IX.P.la
                                                   (p.  47)

                                                   VII.P.Ib
                                                   (P-  44)
                                                              Dilution  Factor =
                                50 ml
                                                                                ml  sample  used in dilutTorf

-------
EFFLUENT MONITORING PROCEDURE:
     Determination of Nitrate-Nitrite  Nitrogen  and  of
     Nitrate Nitrogen, Cadmium  Reduction  Method
OPERATING PROCEDURES
      STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
Q, Calculation of
   mg/liter Nitrate
   Nitrogen in a Sample
   Subtract the mg/liter of
   nitrlte-N in the sample
   from the mg/1iter of
   nitrite-N plus nitrate-N
   In the sample to obtain
   the concentration of
   nitrate-N.
la. Since the procedure measures the total  nitrite
    concentration in a sample, the nitrite  concen-
    tration of samples must be determined with
    reduction and without reduction.   The nitrate
    concentration of a sample is then determined by:

    NO.-N = (NOp+NO^-N) TOTAL - (NO.-N)   WITHOUT
                        WITH             REDUCTION
                        RE-
                        DUC-
                        TION
    These concentrations were recorded on the data
    sheet in Columns (A) and (C) respectively.
                           2, Record the answer in
                              Column (E) on the data
                              sheet provided.
IX.Q.la
(P-  47)
R. Calculation of
   mg/liter Nitrate
   in Sample
1.  Multiply the value found
   for nitrate-nitrogen
   (N03-N) by a factor of

   4.43.

2.  Record the answer in
   Column (F) on the data
   sheet provided.
la. (NC3-N) x (4.43) = mg/liter Nitrate in sample.

Ib. NO,-N value was calculated in Procedure Q and
    recorded in Column (E).
IX.R.lb
(p. 47)
S. Calculation of
   mg/liter Nitrite
   in Samples
1.  Multiply the value found
   for nitrite-nitrogen
   (N02-N) by a factor of

   3.29.


2.  Record the answer in
   Column (G) on the data
   sheet  provided.
la. (N02-N) x (3.29) = ing/liter Nitrite in  sample.

Ib. NO»-N value is found by using the calibration

    curve for non-reduced samples as in Procedure P
    and recorded in Column (C).
IX.S.lb
(p. 47)
                                                                                                            E10.A-39

-------
                                          TRAINING GUIDE

              SECTION                      TOPIC
                  I*              Introduction
                 II               Educational Concepts  -  Mathematics
                III               Educational Concepts  -  Science
                 IV               Educational Concepts  -  Communications
                  V               Field  and Laboratory  Equipment
                 VI*              Field  and Laboratory  Reagents
                VII*              Field  and Laboratory  Analysis
               VIII*              Safety
                 IX*              Records  and Reports
              *Training guide materials are presented here under the headings marked*.
               These standardized headings are used throughout this series of procedures.
Fin A-

-------
EFFLUENT MONITORING PROCEDURE: Determination of Nitrate-Nitrite Nitrogen and Nitrate
                              Nitrogen, Cadmium Reduction Method
INTRODUCTION
     Section
                                 TRAINING GUIDE NOTE
 REFERENCES/RESOURCES
                 The cadmium reduction procedure for nitrate-nitrite
                 nitrogen provides a sensitive method for the deter-
                 mination of nitrate singly, or nitrite and nitrate
                 combined in drinking, surface, and saline waters.
                 The method is commonly used to determine both
                 nitrate-N and nitrite-N in water samples.

                 The procedure described in this EMP is applicable
                 for range of 0.01 to 1.0 ing/liter of nitrate-
                 nitrite nitrogen.  However, the range may be
                 extended by appropriate sample dilution.
                The test described in this instruction can be found
                in the 1979 EPA Methods Manual on p. 353.3-1, entitled1
                Nitrogen, Nitrate-Nitrite (Cadmium Reduction Method).
                Another reference which contains an acceptable
                test for NPDES monitoring is on page 423 of the 14th
                edition of Standard Methods.

                The major sources of nitrogen entering the environ-
                ment are:  through the heavy application of nitrogen-
                ous fertilizers which cause agricultural runoffs,  as
                the end products of aerobic stabilization of organic
                nitrogen, in domestic sewage,  through  animal  and
                plant processing wastes, in animal  manure, through
                the atmosphere and fn various  types of industrial
                effluents.

                While nitrogen is essential  to our  survival  (as in
                the make-up of amino acids and proteins), when  it
                exists  as nitrate and nitrite  it can be toxic.  A
                limit of  10 mg/1  nitrate-N and  1  mg/1  nitrite-N is
                recommended for public  water sources.   The desirable
                criteria  is virtually 0 mg/liter.

                In ruminant animals (i.e.  cows) nitrates may  be
                internally  reduced by bacteria  present in the rumen
                to nitrites.   The nitrites have been found to be tox-
                ic to these animals.  Dr.  Joptha  E.  Campbell,  (Chief,
                Food  Chemistry Unit,  Milk  and  Food  Research,  Environ-
                mental  Sanitation Program, Public Health Service,
                U.S.  Department of H.E.W., Cincinnati,  Ohio,  1968)
                has reported methemoglobfnemia  in cattle receiving
                water containing  2.790  mg/liter of  nitrate.

                Nitrates  in high  concentrations have also been  found
                to stimulate vegetative  growth  under favorable  con-
                ditions.  Heavy undesirable  growth  in  fresh water  can
                lead  to eutrification of important  waterways.
1. Methods for Chemical
   Analysis of Water and
   Wastes, 1979, EPA-
   EHSL,  Cincinnati,
   Ohio 45268, p. 353.3-1.

2. Standard Methods  for the
   Examination of Water and
   Wastewater, 14th  ed.,
   1976,  APHA, New York,
   New York, p.  423.
   Federal  Water Pollution
   Control  Administration
   Water Quality Criteria.
   U.S.  Sovernment Printing
   Office,  Washington,  D.C.
   1968.
                                                                                     E1D.A-41

-------
    EFFLUENT  MONITORING PROCEDURE:  Determination of  Nitrate-Nitrite  Nitrogen  and  Nitrate
                                   Nitrogen,  Cadmium Reduction Method
    FIELD AND  LABORATORY REAGENTS
                                                                               Section  VI
                                     TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
    D.
Samples should be analyzed for nitrate nitrogen as
soon as possible after sampling to avoid any change
in nitrogen balance due to biological activity.  If
analysis can be made within 24 hours, the sample
should be preserved by refrigeration at 4°C.
Samples should be preserved with sulfuric acid if
they are to be held more than 24 hours.  To pre-
serve samples for analysis, add 2.0 ml of con-
centrated sulfuric acid per liter of sample and
store at 4°C.
E10.A-42

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and Nitrate
                                Nitrogen, Cadmium Reduction Method
FIELD AND LABORATORY ANALYSIS
                                                                           Section  yil
                                  TRAINING GUIDE  NOTE
                                                       REFERENCES/RESOURCES
N.3b
  Check the column efficiency when  it  is suspected
  that column efficiency  is decreasing, as  indicated
  by  suspected  low concentration levels.  Prepare
  working standard nitrate solutions,  and pass them
  through the column.   (Begin at E. Preparation of
  Nitrate Working Standards.)  If the  absorbance for
  the known concentration does not  give an average
  between 96% and 104% of your standard curve value
  for reduced nitrate standards of  equivalent concen-
  tration, the column must be reactivated.

  REACTIVATION OF COLUMN

  1. Empty cadmium granules from column into a clean
    beaker.

  2. Wash with distilled water 3 times.

  3. Add enough dilute HC1 to cover granules.

  4. Swirl contents.

  5. Decant HC1.

 6. Wash with distilled water 3 times.

 7. Add 100 ml  CuSO. solution to granules.

  8. Swirl contents of beaker for approximately 5
    minutes  until the blue color fades to colorless.

  9. Decant liquid leaving the granules.

10. Repeat steps 7,  8,  and 9 until  a  very fine
    brown-red precipitate forms.

11. Wash granules with  distilled water (approximately
    10 times) until  precipitate is  removed.

12. Place granules on the 60 mesh sieve.

13. Shake to remove  the small  particles (the particles
    which remain on  the sieve are the ones  you want.)

14. Repack column (packing must be  loose).

15. Activate the column (See F.2).

16. Standard curve using nitrate v/orking  stadards
    must be  re-established.

17. Check column efficiency as  described  in N,
    Checking  Column  Efficiency.
                                                                                     E10.A-43

-------
EFFLUENT MONITORINGPROCEDURE:  Determination  of Nitrate-Nitrite Nitrogen  and  Nitrate
                               Nitrogen,  Cadmium Reduction  Method
FIELD AND LABORATORY ANALYSIS
                                         Section VII
                                  TRAINING  GUIDE  NOTE
                                     REFERENCES/RESOURCES
 H.1.2a
 K.l.Za
 O.lb
 P.lb
                              to
Since a dilution is only part sample, when the
absorbance reading obtained for 1t Is converted
a concentration using a calibration curve, the
concentration obtained 1s only that of the dilution.
To obtain the mg/Hter concentration of the sample,
the mg/liter concentration of the dilution must be
multiplied times the amount of dilution (must be
multiplied times the dilution factor).  For a 1/2
dilution (25 ml sample/50 ml  total volume) the
dilution factor would be 2 (the dilution is only half
sample).  For a 1/5 dilution  (10 ml of sanple/50 ml
total volume) the dilution factor would be 5.  Below
is a table of some dilution factors when the sample
is diluted to a 50 ml volume.
                 ml  of  Sample  per
                 50  ml  Total Volume
                        25
                        10
                         5
                         1
                         0.5
                         0.05
  Amount of
  Dilution

   1/2
   1/5
   1/10
   1/50
   1/100
   1/1000
                               Dilution
                               Factor

                                 2
                                 5
                                 10
                                 50
                                 100
                                 1000
                 The dilution  factor for any dilution may be calcu-
                 lated  by dividing the ml of sample used 1n the
                 dilution into 50:
                 Dilution Factor =
          50 ml
ml sample used in dilution
                 Ex. 2 ml of sample diluted to 50 ml
                    50 -
                The dilution factor for this dilution would be 25.

-------
 EFFLUENT MONITORING PROCEDURE: Determination of Nitrate-Nitrite Nitrogen  and  Nitrate
                               Nitrogen, Cadmium Reduction Method
FIELD AND LABORATORY ANALYSIS
                                                           Section  VII
                                  TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
M.2a
A calibration curve is prepared by plotting the
measured absorbance of each of the workinq

standard versus the concentration in the working
standard as shown below.
                            CONCENTRATION OF N03 or N02 - N. mg/Iiter
                                                                                    E10.A-45

-------
EFFLUENT MONITORING PROCEDURE:
               Determination of Nitrate-Nitrite Nitrogen and Nitrite
               Nitrogen, Cadmium Reduction Method
SAFETY
                                                           Section VIII
                                  TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 C.Z.ld
Cadmium metal is highly toxic thus caution must be
exercised in the use of cidmium.  Cadmium metal
should never be handled directly since cadmium has
been shown to have cumulative effects.  Rubber
gloves should be used whenever cadmium must be
handled.  A mask should be worn during the filing of
cadmium and the filing should be done in a hood.
The waste cadmium should be disposed of in an
appropriate manner which conforms to Federal, State
and local pollution control regulations.
cm fl.

-------
 EFFLUENT MONITORING PROCEDURE: Determination of Nitrate-Nitrite Nitrogen and Nitrate
                               Nitrogen, Cadmium Reduction Method
 RECORDS AND  REPORTS
                                                           Section  IX
                                  TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
L.2.25
H.4a
L.3.2a
M.5a
O.la
P.la
Q.la

R.lb

S.lb
                 You will need the following Key to use the Example
                 Data Sheet found on the next page:
                KEY TO DATA SHEET

(B) Record the absorbances of the column-reduced
    nitrate working standards and of the column-
    reduced sample(s) in Column (B).

(D) Reccrd the absorbances of the non-reduced
    nitrite working standards and of the non-reduced
    sample(s) in Column (D).

(A) Read the mg/liter (concentration) of
    Total NOj+NOj-N in the column-reduced sample(s)

    from the corresponding calibration curve and
    record the answer(s) in Column (A).

(C) Read the mg/Hter (concentration) of NO--N in

    the non-reduced sample(s)  from the corresponding
    calibration curve and record the answer(s) in
    Column (C).

(E) Subtract:   Value (A) - Value (C) = Value (E)

(F) Multiply:   Value (E) x 4.43 =  Value (F)

(G) Multiply:   Value (C) x 3.29 =  Value (G)
                                                                                   E1Q.A-47

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate-Nitrite Nitrogen and Nitrate
                                Nitrogen, Cadmium Reduction Method
RECORDS AND REPORTS
Section IX
                                      EXAMPLE DATA SHEET
See Key on Page No.  7-47
SAMPLE
NUMBER
mg/liter
TOTAL
N02+N03-N
(A)
Reduced Nitrate
Working Standards
2
3
4
5
6
0.05
0.10
0.20
0,50
1,00
Reduced Sample (s)






Non-reduced Nitrite
Working Standards
2
3
4
5
6
\ /
\ /
X
/ \
/ \
Non-reduced Sarnple(s)



\ /
X
/ \
ABSORBANCE
OF TOTAL
N02+NQ3-N
(B)











\ /
\ /
X
/ \
/ \

\ /
X

mg/Hter
N02-N
(C)

\ /
\ /
X
/ \
/ \

\ /
X
/ \

0.05
0.10
0.20
0.50 '
1.00




•J
ABSORBANCE
N02-N
(D)

\ /
\ /
X
/ \
/ \

\ /
X
/ \










mg/liter
N03-N
(E)

0.05
0.10
0.20
0.50
1.00

\ /
X
/ \

\ /
\ /
X
/ \
/ \




mg/ liter
N03
(F)

0.22
0.44
0.89
2.22
4.43

\ /
X
y \

\ /
\ /
X
/ \
/ X




mg/Hter
N02
(G)

\ /
\ /
X
/ X
/ \

\ /
X
/ X

0.16
0.33
0.66
1,65
3.29




   E10.A-48

-------
ABSORBANCE
p p — '
 00 O
o o o
0.40
0.20
i
|
I
0.00"
EFFLUENT MONITORING PROCEDURE: Determination of Nitrate-Ni
. Cadmium Reduction Method
RECORDS AND REPORTS
DETERMINATI
(Reduced Ni
SIGNATURE 0
DATE GRAPH
4 ,{{.,_
.- i. . 4
- *. . i . |
" i ' ! . ; ; i ;
. i . t . - \ '
. i . . ..
* 4 ' ! • I •
. i . . . I . i . .
* . ; - } . i . i
i . - . i * i . j .
: ; i ::::•!
t . i . i . .
trite Nitrogen and of Nitrate Nitrogen,
SECTION
ON OF TOTAL NO^NO^M
trate Standards)
CALIBRATION GRAPH
F PREPARER:
WAS PREPARED
- J - * • t - • •
, i .-,,.,.,,
	
... , i . . .
... , . , i ,
, f » S . . * | .
: . ^ { , ^ i . .
.4 ..4 . - - f - *
- - • • » - i • *
, , | - . . i . t


iiil
. v , i . ,.
,i:|:L






-
-
,




*
k— i






-1
i_


















. i.
•|
, ,
. .
• •
i
i ,
i •
1 •
....
i • i •
! • t •
4 ' ' 1
, . . ,
* f T

f"~
•


-




r

t
_a,

'




.j



*— i


*




1










-

. i . t .









t
i
*~~i
• r ;






;:i
. t
.
:






; : : i ; : :



. . | : i . ...
H~




i
i
i
r
-4—
•


1 ,
.
*
i
J
i
— *—
;
«-


,
' >
•
i
i


~

•
	 i




—
...



t • t
r • i •
• I
. . . ,

T
1 < -t
,
•i;
— •—• t *
T"
-l
-
„.'
*•



-
•
'
1






i
.
;
' '





+ f -*
•

' t
i



f
-
-
•-









IX


-M
.4 -

..... r .
-^•* ' • : J





_
\




-;
r
1
I
-H
. .
1 -
.1
•
i_j
r
1
^1





+ * * -,
•* - f-
••- * — i — j —
: !• • ;
4fF

I*'






—
i .
...
i
i












•

:
r

1 !
— L

"
-.

^
u
f t
-
•
-
1
1
!'







+

f
•f
•
r
ij
i
i
t
•
*
. t-.

. . .
* -i

i
* t
+ -* <
r *
t *
v -
f"
— *— »
* *
1
t •
t
; ;
f
i !
i
id
0.10
0.20
0.30        0.40         0.50        0.60        0.70
    CONCENTRATION  OF NITRATE NITROGEN, mg/liter
                                                                                  0.80
                                                                                  0.90
1.0C
                                                                                                  E10.A-49

-------
                                                                                                                    E10.A-50
    1.00
    0.80
f.
CO
C£
O
1/1

S  0.60
   0.40 —
   0.20
EFFLUENT MONITORING PROCEDURE:
Determination of Nitrate-Nitrite
Cadmium Reduction Method
RECORDS AND REPORTS
DETERMINATION OF HO^-N
(Non-reduced Nitrite Standards)
J CALIBRATION GRAPH
SIGNATURE OF PREPARER:
DATE GRAPH HAS PREPARED
I i 4. 1 . j . 1 . 1 . i . I . | .. .
U - i 1 i » J . I . i . .- .
i . i I i '

. f -T r } , 4 . i | i . ; .
L....f; .}
f - — f * - j-
* - + i -•- » i



1






i




*



.,!....
-;•-:•- J ; r * ; ; ; ; i 	

L i _| . } . . t



i * * * * 1


, ~[ * j • i
. , - *
r i ; ' ' :



*+.{...
. : i . , . ,















-





























*












1 1 :















i 	
j

I











4 *• -k-







• : i : i :_,
* -
1 s |

> i . > ' > *
I ' f ' 1 ! *> i j
j r it
i !* : ; '

T f * * f f - ' '
i 4 . . . | . . ,
>. - - . ,- i 4 ,
L: :;:;::
-_i:l:i:.
i- } : .
f :i ! i : : :

I * 1 . i i j .

1 , . 4 . i -


1 I '11'"







4







'



i












4-
*
f













•














































'






•

















T - f 4 r









1 -









i



















































'



t -
4- -



•


r • j






























,



-
•

*




•



f
1


















^







-




j





















,


<




-


*
_.














:





•
•

•




i













































1 !










t






















•
|











*
















>-


















..,,...
*• • * * r }







*



4



















-




'










"1






t



-










>

•
.
•
.:
. i 1 	


'
_*
_










-





-







-.






r
















•
|
. i
i

\ *



; {




•










.














Nitrogen and of Nitrate Nitrogen,

*












•



•




-











i






t I






1
I
•



-

J












*
.
t
J





"







!






.

-4
SECTION IX
t f

:




-t

-
1
'.
1
t
•
-
-
\
















:.











-






-




-



















-












-






-
-
•
-

-
-


-












-
-









-



-











.




;




i
•


j-
t
i
t



-•



j

^


f
K~*~
i
•



-
--
1
^



-



f
-*
I
\;
t-
4
i™™+_
t
i
0.10
0.20        0.30
                                                    0.40        0.50        0.60        0.

                                              CONCENTRATION  OF  NITRITE  NITROGEN, mg/liter
                                                                          70
                                                                        0.80
0.90
1.00

-------
                          A PROTOTYPE FOR DEVELOPMENT OF
                          ROUTINE OPERATIONAL PROCEDURES

                                      for the

                    DETERMINATION OF NITRATE IN DRINKING WATER
                       AND WASTEWATERS BY THE BRUCINE METHOD
                                   as applied In

                            WATER TREATMENT FACILITIES
                          WASTEWATER TREATMENT FACILITIES
                                    and in the
                         MONITORING OF EFFLUENT WASTEWATERS
              National Training and Operational Technology Center
                       Office  of  Water  Program Operations
                      U.S.  Environmental  Protection Agency
CH.N.n/n.lab.WMP.1.11.77
                                                                         E10.B-1

-------
Effluent Monitoring Procedure:  Determination of Nitrate in Drinking Water
                                and Wastewaters by  the  Brucine Method

1. Analysis Objectives;

   The learner will determine the nitrate content of a  sample of drinking
   water or wastewater effluent.

2. Brief Description of Analysis:

   Brucine sulfate-sulfanilic acid color reagent is added to a series  of
   nitrate standards and to the sample.  The yellow color which  develops
   is read in a spectrophotometer at 410 nm.  A calibration graph is
   prepared, and the nitrate nitrogen content of the sample is determined
   from the graph.

3. Applicability of the Procedure:

   The method works well in waters having salinities which range from
   that of fresh water to sea water.

   a. Range of Concentration:

      The method is recommended for use in the range of
      0.1-2.0 mg of nitrate nitrogen/1.

   b. Pretreatment of Sample:

      Filtration through a 0.45 pm membrane in the
      case of samples which are turbid or contain solids.
      See C.2.

   c. Treatment of Interferences 1n the Sample:

      Chlorine is an interference and is removed by addition
      of sodium arsenite solution.  See C.e.

4. Source of Procedure:

   Standard Methods for the Examination of Water and Wastewater,  14th  ed.,
   pg.  461, Method 213C, 1975.
                                                                  E10.B-3

-------
 EFFLUENT MONITORING PROCEDURE: Determination of Nitrate in Drinking Water and Wastewaters
 i^——-^^—  by the Brucine Method
                                                                                E10.B-4
 OPERATING PROCEDURES
        STEP SEQUENCE
     IMFORMATION/CPr.RATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Equipment Preparation
   1. Cleaning of
      glassware
   2.  Balance Inspection
B.  Reagent Preparation
   1.  Distilled Water
   2.  Stock  Nitrate
      Solution
 1.  Clean all Glassware and
    rinse with  tap and
    distil led water.

 1.  Check all balances for
    cleanliness and proper
    operation.
                                                        V.A.1.1.
                                                        (P. 16)
la.  Consult the manufacturer's manual if the balance
    does not operate properly.
1. Prepare approximately five
   liters of distilled water
   for use in this procedure.
                                                          la,
                                                          Ib.
    Either distill  the water, or obtain distilled
    water from some other source.

    Throughout the  remainder of this procedure, unless
    otherwise stated, the term water means distilled
    water.
1. Add about 500 ml of water
   to a 1 liter volumetric
   flask.

2. Weigh 0.721B g of
   anhydrous potassium nitrate

3. Transfer it to the flask.

4. Swirl the flask.

5. Add water to the 1000 ml
   mark.

6. Thoroughly mix the contents
   of the flask.
la.  Estimate the 500 ml.
                                                          2a,  Use  an  analytical  balance.
                                                          4a.  To  dissolve the  solid.
                                                          6a.  The  concentration  of  this  solution  is  0.1  mg N/ml.

-------
 EFFLUENT MONITORING PRUCEDURF:   Determination  of Nitrate  in  Drinking  Water and  Wastewaters
                                 by  the  Bruclne Method
 OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                      TRAINING
                                                     GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   3.  Standard  Nitrate
      Solution
   4.  Brucine  Sulfate-
      Sulfanilie  acid
      Solution
7. Transfer the solution to a
   1 liter glass stoppered
   bottle.

1. Pipet 10.0 ml of the stock
   nitrate solution into a
   1 liter volumetric flask.

2, Add water to the 1000 ml
   mark.

3. Thoroughly mix the contents
   of the flask.

4. Transfer the solution to a
   1 liter glass stoppered
   bottle.

1. Weigh 1,0 g of brucine
   sulfate.
                           2. Weigh  0.1 g of  sulfanilic
                             acid.

                           3. Measure  70 ml of water.

                           4. Transfer it to  a 250 ml
                             Erlenmeyer flask.

                           5. Heat the water  to almost
                             bo iling.

                           6. Transfer the brucine
                             sulfate and sulfanilic acid
                             to  the flask.
la. Prepare this solution just prior to use.

Ib. Use a volumetric pipet.

2a. Use a trip balance


3a. The concentration of this solution is 1.0 vg N/ml
la.
Caution: this material  is extremely toxic.   If any
is spilled, wipe it up  with damp tissues,  discard
the tissues, and wash your hands thoroughly.
                                                          Ib. Use a trip balance.
                               3a.  Use a  100 ml  graduated cylinder.
                                                                                                          E10.B-5

-------
 EFFLUENT  MONITORING  PROCEDURE:  Determination of Nitrate in Drinking Water and Wastewaters
                                by the Brucine Method
                                                                                                          E10.B-6
 OPERATING  PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
1.  Reagent Preparation
   (Continued)
  5.  Sulfuric acid
     Solution
 7. Measure 3 ml of concentra-
    ted hydrochloric acid, HCl.

 8. Add it to the flask.

 9. Swirl  the flask.

10. Allow the solution to cool
    to room temperature.

11. Measure 30 ml  of water.

12. Transfer it to the flask.

13. Thoroughly mix the contents
    of the flask.

14. Transfer the solution to a
    100 ml glass stoppered
    bottle.

 1. Measure  125 ml  of water.
                          2,  Transfer it to a 1 liter
                             Erlenmeyer flask.

                          3.  Measure 500 ml of
                             concentrated sulfuric acid,
                             H2S04.


                          4.  Pour about 100 ml  of the
                             acid into  the Erlenmeyer
                             flask.
7a.  Use a 10 ml  graduated cylinder.
                                                         9a. To dissolve the solids.
                                                         lla. Use a 100 ml graduated cylinder.
                                                         13a.  Caution: this solution is also extremely toxic.
 la.  Use a  100 ml  graduated  cylinder.
                                 3a.  Use a  500 ml  graduated cylinder.
                                 4a.  Pour it down  the sides  of the  flask.

-------
 EFFLUENT MONITORING PROCEDURE;   Determination of Nitrate in Drinking Water and Wastewaters
              	                 by the Brucine Method
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   6.  Sodium Chloride
      Solution
5. Thoroughly mix the contents
   of the flask.

6. Repeat steps 4 and 5 until
   all  of the acid has been
   added.

7, Allow the solution to cool
   to room temperature.

8. Transfer the solution to a
   1 liter glass stoppered
   bottle.

1. Add about 70 ml of water.
                           2.  Transfer it to a 250 ml
                              Erlenmeyer flask which
                              has a mark at the 100 ml
                              level.

                           3.  Weigh 30 g of sodium
                              chloride, NaCl.

                           4.  Add it to the flask.

                           5.  Swirl the flask.

                           6.  Add water to the 100 ml
                              mark.

                           7,  Thoroughly mix the contents
                              of the flask.
5a. Caution: a large amount of heat will be generated
la. Prepare this solution only if the sample is known
    to contain chloride.  Two ml  of it are needed per
    sample and standard.

lb. Estimate the 70 ml.
                                5a. To dissolve the solid.
                                                                                                           E10.B-7

-------
  E_F_F_LUE_NT_ MUNI TOR ING PROCEDURE:   Determination  of Nitrate  in  Drinking  Water  and Wastewaters
                                  by  the  Brucine  Method
                                                                                E10.B-8
  OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
 B.  Reagent Preparation
    (Continued)
 7.  Sodium  Arsenite
    Solution
C.  Sample Pretreatment
1.  Storage and
   Preservation
8. Transfer the solution to a
   100 ml glass stoppered
   bottle.
1. Add about 50 ml of water to
   a 100 ml volumetric flask.
                            2.  Weigh  0.5  g of  sodium
                               arsenite,  Na
                            3.  Transfer  it  to  the
                               volumetric flask.

                            4.  Swirl the flask.

                            5.  Add water to the  100 ml
                               mark.

                            6.  Thoroughly mix  the contents
                               of the flask.

                           7. Transfer it to a  100 ml
                              <]lass stoppered bottle.
la.  Prepare this solution only if the sample is known
    to contain chlorine.

Ib.  Estimate the 50 ml.
                               2a. Caution: this material is extremely toxic.  If any
                                   is spilled, wipe it up with damp tissues, discard
                                   the tissues, and wash your hands thoroughly.

                               2b. Use a trip balance.
                               4a.  To dissolve the soid.
                               6a.  Caution: This solution is also extremely toxic.
1.  Start the nitrate
   determination promptly
   after sampling.
la.  Standard Methods  offers three preservation
    techniques when needed:

    i)  Concentrated sulfuric acid, 0.8 ml/liter of
       sample.  If this technique is used, the sample
       must be neutralized to a pH of about 7 just

-------
 EFFLUENT MONITORING PROCEDURE:  Determination  of  Nitrate  in  Drinking  Water  and  Wastewaters
-  - --                           by the Brucine Method
 OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
                 TRAINING
                GUIDE  NOTES
C.  Sample Pretreatment
   (Continued)
   2.  Solids,  Turbidity
   3.  Chlorine
                                      before  beginning  the  determination.
                                  ii)  Mercuric  chloride,  54 ing/liter of  sample.
                                 iii)  Chilling  to  just  above the  freezing  point.

                                b.  For later  convenience  in handling the chlorine
                                   interference, filter 1  liter of  sample.
1.  If solids  and/or turbidity
   are present in  the sample,
   filter it  through a 0.45  \im
   membrane filter.

1.  Determine  the  free and  com-
   bined  chlorine  content  of
   the sample in mg/1.
                          2.  For each 0.1  mg of chlorine
                             (free and combined),  add 1
                             drop of the sodium arsenite
                             solution.
   The filter will be plugged quickly by
   sample.  However, only about 20 ml of
   needed for the determination.
                          3.  For each  50  ml  of sample,
                             add 1  drop of  the sodium
                             arsenite  in  excess.
solids in the
filtrate is
  .  For  samples of wastewater  effluent,  two methods  are
    approved:  amperometric  or titrimetric.   For  samples
    of drinking water, one method  is approved:  N,
    N- Diethyl-Phenylenediamine  (DPD).   See the
    appropriate Effluent Monitoring Procedures  (EMP's).

2a.  Example calculation:

    0.1  = mg of  free chlorine  per  liter  of sample

    1.0  = mg of combined chlorine  per  liter of  sample

    Then 1.1  = mg  of free  and  combined chlorine per
    1iter of sample.

    Therefore add  11  drops of  the  sodium arsenite for
    each 1  liter of  sample.  Only  about  10 ml of  sample
    will  actually  be needed  for the determination, but
    it  is convenient to actually collect a larger
    volume such as  1  liter.
                                                                                                          E10.B-9

-------
 _EFFUJIENT_MUMTOR1N6 PROCEDURE:   Determination of Nitrate in Drinking Water and Wastewaters
                                 by the Brucine Method
                                                                                                          E10.B-10
 OPERATING PROCEDURES
         STEP SEQUENCE
                                     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Sample Pretreatment
   (Continued)
D.  Procedure
   1.  Sample
4.  Thoroughly mix the contents
   of the sample container.
1. Assemble 9 large test tubes
   in a wire rack.

2. Pipet 1.0 ml  of sample
   into a large test tube,

3. Pipet 9,0 ml  of water into
   the same tube.

4. Mix the contents of the
   tube.

5. Pipet 5.0 ml  of the same
   sample into a second large
   test tube.

6. Pipet 5,0 ml  of water into
   the same tube.

7. Mix the contents of the
   tube.

:. Pipet 10.0  ml  of the same
   sample into a  third large
   test tube.
                                la.  There must be an empty rack space next to each
                                    tube.

                                2a.  Use  a graduated  pipet.
                                                          3a.  Use a  graduated  pipet.
                                                          4a.  This  is  a  10%  sample  dilution.
                                                          5a.  Use  the  same  pipet  as  in  2a.
                                                          6a.  Use  the  same  pipet  as  in  3a.
                                                          7a. This  is a  50% sample dilution.
                                                         8a. Use  the same pipet  as  in  2a.
                                                         8b. This  is an undiluted,  or  100%,  sample.
                                                         Be. Use  a clean graduated  pipet  to  measure  volumes  of
                                                             different^ samples.
                                                         8d. The  idea of preparing  three  sample dilutions  is
                                                             that one of them will  give a  result within  the
                                                             applicable range of the test.   Once experience
                                                             with the sample source is gained, it will not be
                                                             necessary to do more than one sample dilytion.

-------
EFFLUENL MU NIT C RIM G PRUlEDU R F:   Determination of Nitrate in Drinking Water and Wastewaters
              	                 by the Brucine Method
OPERATING PROCEDURES
D. Procedure (Continued)
2. Standards














3. Sodium Chloride







STEP SEQUENCE

1. Pipet standard nitrate
solution (B.3) and water
into the six remaining
large test tubes.











1, Place the rack in a tray
containing tap water or
cold water from a fountain.
2, Add about 6 ice cubes to
the tray.
3. Pipet 2.0 ml of the sodium
chloride solution into each
of the 9 tubes.
INFORMATION/OPERATING GOALS/SPECIFICATIONS

la. So as to obtain the following solutions:


Cone, of
Nitrate
Tube ml of Standard ml of Nitrogen in
Number Nitrate Solution Water mg/1
1 0.0 10.0 0.0
2 1.0 9.0 0.1
3 2.0 8.0 0.2
4 4.0 6.0 0.4
5 7.0 3.0 0.7
6 10.0 0.0 1.0
Ib. Use one 10 ml graduated pipet for the standard
nitrate, and a second for the water.
la. The water level should be about half-way up the
tubes.

2a. If cold water is used, step 2 is unnecessary.

3a. Use a 10 ml graduated pipet only if the sample
contains Cl~.

TRAINING
GUIDE NOTES
























                                                                                                          E10.B-11

-------
 EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate in Drinking Water and Wastewaters
 ~™—•——^—^——^—^^^^~~   by the Brucine Method
                                                                                                          E10.B-12
 OPERATING PROCEDURES
D. Procedure (Continued)
   4.  Sulfuric Acid
   5.  Turbidity
        STEP SEQUENCE
4. Thoroughly mix the contents
   of each of the 9 tubes.

1. Pipet 10.0 ml of the sul-
   furic acid solution into
   each of the 9 tubes.

2. Thoroughly mix the con-
   tents of each of the
   9 tubes.

3. Allow the tube contents to
   cool  to room temperature.
1.  Check the three sample
   tubes for turbidity.
                          2. Turn on  the Spec  20 and
                             allow  it to warm  up.

                          3. "Zero" the instrument
                             using  tube number 1.

                          4. Measure and record the
                             absorbancies of those
                             sample tubes whose contents
                             are turbid.
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
4a. By hand swirling.
4b. Do not use a vortex-type mixer.

la. The rack is still in the cold water.
Ib. Use a 10 ml graduated pipet.
                                                           2a.  By hand  swirl ing.
                                                           2b.  Do not  use a  vortex-type mixer.
3a. The ice cubes or cold water will probably have to
    be replaced.
3b. Do this step only if the sample is known to con-
    tain chloride.

la. If more than one type of sample is being run,
    there may be more than three sample tubes.   Check
    all the sample tubes for turbidity.
Ib. If no_ turbidity is present, proceed to D.6,
    Color Development.
Ic. If turbidity j_s_ present, do steps 2, 3, and 4
    below.
                                3a.  At 410  urn.
 TRAINING
GUIDE NOTES

-------
 EFFLUENT MONITORING PROCLDURF:   Determination of Nitrate in Drinking Water and Wastewaters
                                 by the Brucine Method
 OPERATING PROCEDURES
        STEP SEQUENCE
    ENFORMATION/OFFRATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
0.  Procedure (Continued)
   6.  Color Development
   7.  Color Measurement
1, Pipet C.5 ml of the bru-
   cine sulfate-sulfani1ic
   acid reagent into each of
   the 9 tubes.

2. Thoroughly mix the con-
   tents of each of the 9
   tubes.

3. Place the rack of tubes in
   the hot water bath.

4. If the Spec 20 has not been
   used prior to this step,
   turn it on now and allow
   it to warm up.

5. After exactly 20 minutes
   remove the rack from the
   hot water and place it
   back into the cold water.

1. Remove the rack from the
   cold water bath.

2. Dry the outside of the
   large test tubes.

3. Zero the Spec 20 using
   tube 1  as a "blank."

4. Record the absorbanices for
   tubes 2 through 6.

5. Record the absorbancies
   for all sample tubes.
la.  The rack is still  in the cold water,
Ib.  Use a 10 ml graduated pipet.
                                                           2a.  By hand swirling.
                                                           2b.  Do not use a vortex-type mixer.
                                                           3a.  Ninety-five degrees C.
                                                           5a.  The ice cubes or cold water may have to be
                                                               replaced.
                                                           5b.  The contents  of the tubes  should be at room
                                                               temperature.
                                                           4a.  These  six  tubes  are  the  standards  to  be  used  in
                                                               preparing  the  calibration  graph.
                                                                                                           E10.B-13

-------
 EFFLUENT MONITORING PROCEDURE:
      Determination of Nitrate in Drinking Water and Wastewaters
      by the Brucine Method
                                            E10.B-14
 OPERATING PROCEDURES
        STEP SEQUENCE
INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E.  Calculations
1. Prepare a calibration
   graph using the absorban-
   cies and concentrations
   for tubes 1 through 6.

2. If any absorbancies due
   to turbidity were recorded
   for any of the sample
   tubes, subtract those
   absorbancies from the
   appropriate absorbancies
   obtained in D.7.5 above.

3. Determine the mg of
   nitrate nitrogen per liter
   of sample for each
   different sample.

-------
 EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate in Drinking
                                Water and Wastewaters by the Brucine Method
                                 TRAINING GUIDE

SECTION                              TOPIC
   I                       Introduction
  II                       Educational Concepts - Mathematics
 III                       Educational Concepts - Science
  IV                       Educational Concepts - Communications
   V*                      Field and Laboratory Equipment
  VI                       Field and Laboratory Reagents
 VII                       Field and Laboratory Analysis
VIII                       Safety
  IX                       Records and Reports
*Training Guide materials are presented here under the headings marked*.
These standardized headings are used through this series of proceudres.
                                                                        E10.B-15

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Nitrate in Drinking Water and Wastewaters
                             •         gruci-ne Method
FIELD AND LABORATORY EQUIPMENT
                                                          Section  V
                                  TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
   A.I.I
If the glassware is especially dirty and cannot be
cleaned with ordinary detergents, chromic acid
cleaning may be required.

1. Pour 35 ml of distilled water in a 250 ml  beaker.
                 2. Add about  1/8 teaspoon  (simply estimate this
                    quantity)  of sodium dichromate, Ha-Cr^O^. to
                    the water.

                 3. Swirl the  beaker  until  the sodium dichromate has
                    dissolved.

                 4. Keep  repeating  steps  2  and 3  until  no more
                    sodium  dichromate will  dissolve.

                 5. Pour  the  solution into  a  2 liter beaker.

                 6. Slowly  pour 1 liter of  concentrated sulfuric
                    acid, hLSO^,  into the 2 liter beaker.

                    Caution:   Use eyeglasses  and
                               protective  clothing.

                 7. Stir  the  mixture  thoroughly.

                 8. Store it  in a glass  stoppered bottle.

                  9. The cleaning solution should  be  at  a temperature
                    of about  50 C when  it is used.

                 10.  It may  therefore  be  necessary to warm the
                     cleaning  solution,

                 11.  When  using the  warm  cleaning  solution,  fill  the
                     piece of  glassware with the  solution.

                 12.  Allow it  to soak  for 2-3 minutes (or longer).

                 13.  Pour the  cleaning solution back  into the storage
                     bottle.

                 14.  Rinse the piece of glassware ten times with
                     tap water.

                 15,  The  cleaning solution may be reused until it
                     turns green.

                 16,  It should then be discarded.
                                                                      14th Standard Methods,
                                                                      p. 336, section 2.c,2)
    E10.B-16

-------
                          A PROTOTYPE FOR DEVELOPMENT OF
                          ROUTINE OPERATIONAL PROCEDURES

                                     for the

                      DETERMINATION OF FLUORIDE IN POTABLE AND
                               WASTEWATERS USING THE
                            SPADNS COLORIMETRIC PROCEDURE
                                    as  applled in

                             WATER TREATMENT FACILITIES
                           WASTEWATER TREATMENT FACILITIES
                                     and in  the
                          MONITORING OF EFFLUENT WASTEWATERS
               National  Training  and  Operational  Technology  Center
                      Municipal  Permits  and  Operations  Division
                          Office  of Water  Program Operations
                        U.S.  Environmental  Protection Agency
CH.HAL.f.lab.WMP.1.11.77
                                                                         Ell.A-1

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Fluoride in Potable
                                and Waste Waters Using the SPADNS
                                Colorimetric Procedure

1. Analysis Objectives:

   The learner will determine the fluoride content of a water or wastewater
   sample using the SPADNS Colorimetric Procedure.

2. Brief Description of Analysis:

   After distillation to remove interferences the sample is treated with the
   SPADNS reagent.  The amount of color remaining after the reaction of the
   fluoride and reagents is read on a spectrophotometer.  This reading is a
   function of the fluoride concentration.

3. Applicability of this procedure:

   a. Range of Concentration: from 0.05 to 1.4 mg F liter.
   b. Pretreatment of Sample:

   The Bellack distillation procedure must be carried out on all  samples before
   determination by the SPADNS procedure.  The method is applicable to the
   measurement of fluoride in drinking, surface, and saline waters, domestic
   and industrial wastes.   The distillation procedure is covered in a separate
   EMP.

   c. Treatment of Interferences in the Sample:

   The distillation will remove all interferences when carried out properly.
   For a list of interferences consult the training guide notes (VII.B.2.a).

4. Source of Procedure:

   Standard Methods for the Examination of Water and Wastewater;  14th
   Edition 1975.  pg 393.
                                                                   Ell.A-3

-------
                                    FLUORIDE ANALYSIS FLOW SHEET
                                               L
                                SAMPLE
                                   Potable Water
                                        Waste Water

                          DISTILL
ELECTRODE

METHOD
SPADNS

COLORIMETRIC
METHOD
SPADNS

COLORIMETRIC
METHOD
                                                      DISTILL
ELECTRODE

METHOD
AUTOMATED

COMPLEXONE
METHOD

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Fluoride in Potable
                                and Waste Waters Using the SPADNS
                                Coloritnetric Procedure

Equipment and Supply Requirements

A. Capital Equipment;

   1. Spectrophotometer, for use at 570 nm providing a light path of
      at least 1 cm. with cells.
   2. Analytical balance; capable of weighing to 0.1 mg
   3. Still or Ion Exchange column or other source of distilled water
   4. Still for distilling sample - see EMP on Fluoride Distillation
   5. Trip balance - 500 gram capacity

B. Reusable Supplies:

   1. Beakers - 500 ml  glass - 1 each
   2. Flasks - Erlenmeyer - 125 ml - 8 each
   3. Flasks - Erlenmeyer - 500 ml - 1 each
   4. Flasks, volumetric - 50 ml - 8 each
   5. Flasks, volumetric 100 ml - 2 each
   6. Flasks, volumetric 1000 ml - 2 each
   7. Graduated Cylinder - 500 ml - 1  each
   8, Pipets, volumetric - 10 ml - 3 each
   9. Pipets, volumetric - 50 ml - 8 each
  10. Powder funnel - 1 each
  11. Saftey glasses -  1 pair
  12. Spatula - 1  each
  13. Thermometer  0-100 C - 1 each
  14. Wash bottle  - plastic - 1 each

C. Consumable Supplies

   1. SPADNS - Reagent  4,5-dihydroxy-3 [(p-sulfophenyl)azo-2.7 napthalene
      disulfonic acid,  trisodium salt.

      Baker Cat No. 5189   —   10 grams

      Eastman Cat  No. 7309 —   25 grams

   2. Zirconyl chloride - Reagent ZrOCl-

      Baker Cat No. X720   —   500 grams

      Fisher Cat No. Z-80   —   1 Ib.

   3. Sodium Fluoride-NaF-Reagent

      Baker Cat No. 3688   —   1 Ib.

      Fisher Cat No. 5299  —   1/4 Ib.

                or

      Sodium Fluoride Stock Solution
                                                                   Ell.A-5

-------
         EFFLUENT  MONITORING  PROCEDURE:  Determination of  Fluoride in Potable
                                        and Waste Waters  Using the SPADNS
                                        Colon'metric Procedure

         C.  Consumable  Supplies  (continued)

            3.  Orion  Research Inc.  Cat No. 94-06-07

               Hach Chemical  Co.

            4.  Sodium Arsem'te NaAsOp Reagent

               Baker  Cat  No.  3487    —    1 Ib.

               Fisher Cat No. S-225 —    1 Ib.

            5.  Hydrochloric acid HC1 Reagent

               Baker  Cat  No.  9535    —    1 pt

               Fisher Cat No. A-144 -—    1 pt

            6.  Weighing boats-plastic disposable  - 60 each
            7.  Pen or Pencil
            8.  Notebook

         D.  Addresses of  Suppliers  Mentioned

            J.  T.  Baker,  Chemical Co.
            1  Public  Square
            Cleveland OH  44113

            Eastman Organic Chemicals
            Eastman Kodak Co.
            1187 Ridge  Road W.
            Rochester,  NY 14650

            Fisher Scientic Co.
            5481 Creek  Rd.
            Cincinnati, OH 45242

            Hach Chemical  Co.
            P.  0.  Box 907
            Ames,Iowa 50010

            Orion  Research Inc.
            380 Putnam  Ave,
            Cambridge,  MA 02139
Ell.A-6

-------
EFFLUENT MONITORING PROCEDURE:  Determination of Fluoride  Using  the  SPADNS  Method
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE  NOTES
A. Sample Collection
1,  Collect a  minimum of 300 ml
   in a plastic or hard qlass
   container.
B. Sample pretreatment
1.  Add 1  drop (0.05nl)  sodium
   arsenite (NaAs02)  solution
   for every 0.1  mg residual
   Cl.

2.  Distill  the sample.
C. Equipment Preparation

   1. Glassware
   2. Spectroohotometer
      Insoection
1.  Rinse all glassware
   immediately after use with
   taowater.

2,  Rinse several times with
   distilled water.

1.  Clean spectroohotometer.
la
                                                          Ib.
                                                          Ic
                                                          Id,
la
                                                          Ib.
                                                          2a,
                                                          2b.
                                                          2c.
Do not rinse the container in tap water for its
final  rinse.  Tap water usually contains some
fluoride even if the source is not fluoridating.
For distillation a volume of 300 ml  is required.
For the SPADNS procedure 50 ml of the distillate
is used.
No special requirements are necessary for
preservation.
Polyethylene bottles are preferred,  hard glass
(Pyrex, Kimax) is acceptable
Chlorine will bleach the color and is therefore
a definite interference.
Caution:  This solution (Reagent D.6.) is toxic.
Take care not to injest any.

For total or total  dissolved Fluoride the sample
must be distilled.
If interferences are present the sample must be
distilled.
The distillation procedure is covered under a
separate EMP.
la. If detergent is used care should be taken to
    rinse thoroughly to remove any phosphate,


2a. All tap water contains traces of Fluoride and
    could also contain about 1.Omg/1 if fluoridated.

la. Free of dust and dirt.
Ib. Consult manufacturer's instructions.
                                                      i.e.2
                                                      (p. 18)
                                                       VII B.2a
                                                      (p. 20)
                                                                                                         Ell.A-7

-------
EFFLUENT MONITORING PROCEDURE:  Determination  of  Fluoride  Usinq  the  SPAONS  Method
                                                                                                        Ell.A-8
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Reagent Preparation

   1.  Distilled Water
   2.  SPADNS Reagent
1,  Prepare about six liters  of
   distilled water.
   This water should be free
   from fluoride.

1.  Weigh out 0.958q  of SPADNS
   dye.
                          2.  Add  about  200  ml  of
                             distilled  water  to a
                             1  liter  volumetric flask.

                          3.  Transfer the SPADNS dye  to
                             the  1  liter volumetric
                             flask, using a powder
                             funnel.

                          4.  Dissolve the SPADNS dye
                             in distil led water.

                          5.  Weigh  out  0.133  gm
                             zirconyl chloride
                             octahydrate.

                          6.  Dissolve in about 200 ml
                             distil led  water.
                          7.  Transfer  the airconyl
                             chloride  to the  1  liter
                             flask with the SPADNS dye.
la.  Use a still  or pass tap water through an ion-
    exchange column containing a strongly acidic
    cation exchange resin mixed with a strongly
    basic anion  exchange resin.

la.  i.e.,  2.(para sulfophenylazo) -1,8-dihydroxy
    -3,6- nanhthalene disulfonate, also called
    45 dihydroxy -3-(parasulfophenylazo) -27-
    napthalene disulfonic acid trisodium salt.
Ib.  Use an analytical balance and a  plastic weighing
    boat.
                                3a.  Use  a  wash  bottle  to  wash  the  solid  into  the
                                    flask.   The weighing  boat  should  be  washed
                                    three  times and  the washings added to  the flask.
                                5a.  Use  a  plastic  weighing  boat  on  the  analytical
                                    balance.
                                6a.  A  500 ml  beaker may  be  used.
                                6b.  Use  a wash  bottle  to wash  the  solid  into  the
                                    flask.   The weighing boat  should  be  washed  three
                                    times and the  washings  added to  the  flask.

                                7a.  Care should be taken to rinse  the beaker  with
                                    distilled water.

-------
LFFLULNT MONITORING PROCEDURE:   Determination of  Fluoride  Using  the  SPADNS Method
OPERATING PROCEDURES
         STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
2.  SPADNS Reagent
   (Continued)
3. Sodium Fluoride
   (NaF) stock
   solution - 100mq,F~/l
    Add 350  ml  concentrated
    hydrochloric  acid  (HC1),
 9.  Mix well.

10.  Dilute to  the  1  liter  mark
    with distilled water.

11.  Mix well.

1?.  Label.

 1,  Weigh out  0.2210 grams of
    sodium fluoride.
                          2.  Add about 500  ml  of  dis-
                             tilled water to a 1  liter
                             volumetric flask.

                          3,  Transfer the solid to the
                             1  liter volumetric flask
                             using a powder funnel.

                          4.  Use a wash bottle to wash
                             solid into flask.

                          5.  Dissolve the solid.

                          6.  Dilute to the  mark and
                             mix thoroughly.
                           '.  Label
8a.  Caution:  When using the concentrated acid, use
    eye protection.
8b.  Solution may increase in temperature, cool  to
    room temperature before proceeding.
                                                         lla.  Solution  stable  for  at  least  2 years.
la. Use an anlytical balance
Ib. Use a plastic weighing boat.
Ic. Solution can be purchased from Orion Research
    Inc; Cat No. 94-06-07 or Hach Chenieal  Co.;
    Cat No. 232-11.
                                 4a.  The  weighing  boat  should  be  washed  three  times
                                     and  the  washings added  to the  flask.
                                 6a.  Solution  contains  0.1  mg  F  /l.O  ml  (i.e.  100  ppm
                                     n,
                                 6b.  Keep  in a plastic  container.   Solution  is stable
                                     for at  least  6  months.
                                                                                                          Ell.A-9

-------
 EFFLUENT MONITORING PRQCCDURF:   Determination  of  Fluoride  Using  the  SPADNS  Method
                                                                                                         Ell.A-10
 OPERATING PROCEDURES
         STEP  SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
 4.  Sodium  Fluoride
    (NaF) Solution
    10 mg F"/l.
5. Sodium Fluoride
   (NaF) Solution
   1.0 mg F"/liter
6.  Sodium Arsenite
   (NaAsO,,) Solution.
 1.  Add  approximately  50 ml  of
    distilled water  to  a 100
    ml volumetric  flask.

 2.  Pipet  10 ml of the  stock
    fluoride solution  (reagent
    3) into the flask.

 3.  Dilute with distilled water
    to the mark.

 4.  Mix  thoroughly.
 5. Label

 1. Add approximately 50 ml of
   distilled water to a 100 ml
   volumetric flask.

 2. Pipet 10 ml of the fluoride
   solution (reagent 4) into
   the flask.

 3. Dilute with distilled water
   to the mark.

 4. Mix thoroughly.

 5. Label

 1. Weigh out 2.5 grams of
   sodium arsenite.

2. Transfer the  solid to a
   500 ml Erlenmeyer flask.
                                                          2a. Use a  10 ml volumetric pipet.
                                                          3a. Solution contains 0.01 mg  F/l.0 ml  (i.e. 10 ppm),
                                                          4a. Keep in plastic container.
                                                          4b. Solution stable for at least 6 months.
                                                          2a. Use a 10 ml volumetric pipet.
                                                          3a. Solution contains 0.001 mg F~ per 1.0 ml (i.e.
                                                              1.0 ppm F~).

                                                          4a. Solution should be prepared fresh daily.
la.  A trip balance can be used.

-------
EFFLUENT KUNITQRING PROCEDURE:   Determination of Fluoride Using the SPADNS Method
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORKATION/OPERA~ING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
6.  Sodium Arsenite
   (NaAs02) Solution
   (Continued)
E.  Calibration
      By use of a
      standard curve.
3.  Measure out 500 ml
   distil led water.

4.  Dissolve the sodium
   arsenite in about half of
   the distil led water.

5.  Add the rest of the 500 nil
   portion.
3a. Use a graduated cylinder.
1.  Turn on spectrophotometer
   and allow to warm up.

2.  Calibrate the
   spectroohotometer.
                             Preoare a series of
                             standard solutions  usinq
                             the lOmq F'/liter sodium
                             fluoride solution (Reagent
                             D.4).
la.  Refer to manufacturer's manual  for  warm up  time.
    However usually one-half  hour is sufficient

2a.  This can be carried out in two ways;   by use of a
    calibration qraoh or by calculation of the un-
    known.   Section E.I covers the calibration graph
    and Section E.2 covers the calculation.

3a.  The series of standards should be prepared in
    50 ml  volumetric flasks by pipetting  with a
    graduated pioet the indicated amount  of Reagent
    D.4 into a 50 ml volumetric flask and diluting
    to the  volume mark,
                                                                                                                 VII  E.I
                                                                                                                 (p.  21)
                                                                                                          Ell.A-11

-------
EFFLUENT MONITORING PROCEDURF:  Determination o* Fluoride Using the SPADNS Method
                                                                                                        E11.A-12
OPERATING PROCEDURES
STEP SEQUENCE
INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
£. Calibration
   (Continued)
                        3b.  The  following  table  can  be  used  to  prepare  a
                            series  of  standards  containing from 0  to  1.4
                            mg F~/liter.
                                                              ml  of  Reaqent  D.4
                                                                0.00
                                                                1.00
                                                                2.00
                                                                3.00
                                                                4.00
                                                                5.00
                                                                6,00
                                                                7.00
                                                     Cone, when  diluted  to
                                                     50 ml_,_ mg F /11 ter_
                                                          0.00
                                                          0.20
                                                          0.40
                                                          0.60
                                                          0.80
                                                          1.00
                                                          1.20
                                                          1.40
                          .  Transfer the entire 50 ml  of
                            the standards just prepared
                            into 125 ml  Erlenmeyer
                            flasks.
                         5.  Add 10 ml  SPADNS Reagent
                            (Reagent D.2.).
                         6.  Mix thoroughly.
                         7.  Determine  the  absorbance  at
                            570  nm.
                       4a.  All  eight standards can be prepared at the same
                           time.
                       5a.  Use a 10 ml  volumetric  pipet
                       5b.  Caution:  This volume is  critical;  take care to
                           add exactly  10 ml.
                       6a.  Unless  nixed,  the  solution  can layer giving  false
                           readings.
                       7a.  The manufaturer's  manual  on operation of the
                           soectrophotometer  should  be consulted for proper
                           instrument operation.
                                                                                                                m.E.i.
                                                                                                                (p.  22}
                                                          5b
                                                  V.E.1.7a.
                                                  (p.  19)

-------
EFFLUENT HUUITORING PROCEDURE:  Determination of Fluoride Using the SPADES Method
OPERATING PRCCEDURLS
         STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E.  Calibration
   (Continued)
2.  Bv Calculation
 8,  Prepare a  calibration
    curve.


 9.  Analyze the sample

10.  Obtain  concentration usinq
    the graph  prepared here.

 1.  Turn on spectrophotometer
    and allow  to warm up.

 2.  Adjust  the wavelength to
    570 nm.

 3.  Prepare two standards.
                          4.  Pipet 50 ml  of each sample
                             to be run into a series of
                             125 ml  Erlenmeyer flasks.

                          5.  Pipet 10 ml  of SPADN'S
                             Reagent to all flasks.

                          6.  Mix al1 thoroughly.

                          7.  With nothing in the meter
                             adjust  the left side of the
                             absorbance scale to read
                             its maximum,

                          8.  Fill a  sample cell  with
                             0 mg F  /liter standard.
                                                          8a.  Use the attached graph paper.
                                                          8b.  Refer to:  Effluent Monitoring  Procedure:  Prepara-
                                                              tion of Calibration Graphs.
                                                          9a.  Operating  Procedure F.
                                                       V.E.I.8

                                                       IX.E-l.8a
                                                       (p.  19)
3a. Prepare as follows: 0 mg F"/liter (pioet 50 ml  of
    distilled water into a 125 ml  Erlenmeyer flask):
    1  mq F~/liter (pipet 50 ml  of reagent D.5 into  a
    125 nl  Erlenmeyer flask).

4a. Label  each flask with sannle identification.
                                 5a.  This  includes  both standards  (including 0 mg
                                     F /I)  and  samp!es,
                                 7a,  This  is  the maximum (marked as  infinity <»).
                                                                                                           Ell.A-13

-------
EFFLUENT MONITORING PRUCEDURF:   Determination  of  Fluoride Using  the SPADNS Method
                                                                                                        Ell.A-14
OPERATING PROCEDURES
STEP SEQUENCE
INFORMATION/OPERATING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
   2.  By Calculation
      (Continued)
                          9.  Place  the cell  in  the cell
                             compartment,

                         10.  Adjust  the  slit  control  so
                             that a  reading of  0.5 is
                             obtained on  the  absorbance
                             scale.

                         11.  Fill a  sample cell with
                             1.0 mg  F~/liter  standard.

                         12.  Place  in the cell
                             compartment.

                         13.  Read the absorbance value.
                         14. Read all samples without
                            changing any adjustment
                            controls.

                         15. Calculate the concentration
                            of the unknown.
                       lOa.
This is to set the portion of the scale to be
used.
                       13a,
This reading should read about 0.25 to 0.27.  If
this reading is not obtained the make up of the
reagent and standard used should be checked.
                                                        13b. Use Operating Procedure F
                       15a.  Use  the  equation:   X  =
                                                      "
                                                      __
                                                   AQ  -  A,
                                                             X = concentration of the unknown in mg F"/liter

                                                             A = Absorbance reading of the 0 mgF"/l.

                                                             This is set in steo 10 at 0.5
VII.E.Z.lOa.
(p.  22)

-------
EFFLUENT MONITORING PROCEDURE.:  Determination of Fluoride Using  the SPADNS Method
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORKATION/OPCSATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
?.  By Calculation
   (Continued)
F. Procedure
1.  After calibration  the
   samples are tested as
   follows.
                               15a,  (continued)

                                    A.  =  Absorbance  reading  of  the  1.0 mg  F~/liter
                                    (usually  0.25  to 0,27)

                                    Ay  =  Absorbance  reading  of  the  unknown.

                               15b.  Example:   If the sample  absorbance value
                                    obtained  from  the spectrophotometer was 0.32
                                                              X =
                                                                  then

                                                                  0.5-0,32
                                                   0.18
                                                                  0.5-0.25 " 0.25

                                                              X = 0,72 ng F~/liter

                                                              If more than one sample were run the only number
                                                              that would change would be the 0.32.  A
                                                              calculation must be performed for each sample.
                          2. Add 50 ml of satrple to a
                             125 rr] Erlenmeyer flask.

                          3. Add 10 ml SPADNS Reagent,

                          4. Mix thoroughly.
la.  If the calibration graph was used (E.I.), two
    standards should be used each time to verify that
    the calibration curve is still  valid. (Al.O mg F~
    /liter and a 0.2 mg F~/liter).

Ib.  If the calculation procedure was used, the 0 and
    1.0 oom standards rust be run with each batch of
    samples run.

?a.  Use a 50 rrl volumetric pipet.
                                3a.  Use a 10 ml  volumetric pipet.
                                                                                                          Ell.A-15

-------
EFFLUENT MONITORING PROCEDURE.:  Determination  of  Fluoride  Usinq  the  SPADNS  Method
                                                                                                        Ell.A-16
OPERATING PROCEDURES
STEP SEQUENCE
INFORMATION/CREATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Procedure (Continued)
                          5.  Read  the  absorbance  value
                             at  570  nm.

                          6.  Dilute  sample  if
                             concentration  is  not be-
                             tween 0.0 to 1.4  mq
                             F"/liter.
                        6a.  If  the  calibration  graph  is  used.the absorbance
                            value for  a  concentration of 1.4 mg F~/liter has
                            been ploted  do  not  use absorbance  values
                            corresponding  to  values greater than 1.4 mg
                            F"/liter.

                        6b.  If  the  calculation  procedure is used, obtain the
                            absorbance value  and  calculate if  the value
                            calculated is  above 1.4 mg F~/liter dilute the
                            sample.

-------
                            TRAINING GUIDE

SECTION                     TOPIC
   I*               Introduction
  II                Educational Concepts - Mathematics
 III                Educational Concepts - Science
  IV                Educational Concepts - Communications
   V*               Field and Laboratory Equipment
  VI                Field and Laboratory Reagents
 VII*               Field and Laboratory Analysis
VIII                Safety
  IX*               Records and Reports
*Training Guide materials are presented under the heading marked *.   These
standardized headings are used throughout this series of procedures.
                                                                    Ell.A-17

-------
EFFLUENT MONITORING PROCEDURE:
                Determination of Fluoride Using the
                SPADNS Method
 INTRODUCTION
                                                                           Section
                                  TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 I.
     The basis for the SPADNS procedure is the
reaction between zirconium and the SPADNS dye
(SPADNS is an abbreviation of sodium 2-(p-sulfo~
phenylazo)-!,8-dihydroxy 3.6-napthlene disulfonate).
The color of the reaction mixture (water sample plus
reagent) varies from very deep red in the absence of
fluoride  to light red when the concentration of
fluoride is high.   The change in color caused by
small changes in fluoride concentration is not
discernable by eye, but can be readily detected by a
photometric instrument.  The better the instrument,
the better the sensitivity to small  changes of
f1uoride.

     Distillation is not required if comparability
data on representative effluent samples are on file
to show that this preliminary distillation step is
not necessary.  However, manual distillation will be
required to resolve any controversies.

     For drinking water samples analysis to comply
with requirements as listed in the National Interim
Primary Drinking Water Regulations (F. R. Part IV,
12/24/?5-para. 141.23f,10-pg59573\ distillation is
necessary.

     Distillation of the calibration standards is
not necessary should sample require distillation.
    Ell.A-18

-------
EFFLUENT MONITORING PROCEDURE:
                Determination of Fluoride Using the
                SPAONS Method
 FIELD AND LABORATORY EQUIPMENT
                                                                           Section
                                  TRAINING  GUIDE  NOTE
                                                        REFERENCES/RESOURCES
  E.1.7a
  E.l.i
     Should the analyst be using a Bausch and lornb
Spectronic "20" Spectrophtometer there is an EMP,
"Use of a Spectrophototneter" availble.  It would be
of value to the analyst to consult this procedure,

     There is an EMP "Preparation of Calibration
Graphs" that would be of value to the student.
                                                                                Ell.A-19

-------
EFFLUENT MONITORING PROCEDURE:
                Determination of Fluoride Using the
                SPADNS Method
  FIELD AND LABORATORY ANALYSIS
                                                                           Section
                                                                    VII
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
    ,2.a
     Other ions, particularly phosphate, iron and
aluminum can cause significant errors.  The table
below gives some indication of this error.   Should
any of these interferences be present in the sample
at concentrations that will cause error, the sample
must be distilled.  The numbers given show  the
concentration in mg/liter that will cause an error
of plus or minus 0.1 mg/liter at 1.0 mg F*/l.
Substance      Cone
                                                   Type Error
                Alkalinity
                (CaC03)

                Aluminum
                (Al + )

                Chloride
                (Cl")

                Iron,
                (Fe+3)

                Hexametaphosphate
                (CNaP03]g)
                Phosphate
                Sul
                Chlorine
                Color &
                Turbidity
                    5000


                     0.1"


                    7000


                      10


                     1.0


                      16


                     200
                Must be completely
                Removed with Arsenite

                Must be Removed or
                Compensated for.
                                                      Fluoride Determination in
                                                      Water, E.P.A.,  Training
                                                      Manual NTOTC, Cincinnati,
                                                      OH 45258.
                *Above figure is for immediate reading.   Allowed to
                stand two hours tolerance is 3.0 mg/liter, four hours
                tolerance is 30 mg/liter.
    Ell.A-20

-------
EFFLUENT MONITORING  PROCEDURE:
                Determination of Fluoride Using the
                SPADNS Method
  FIELD AND LABORATORY ANALYSIS
                                                           Section
VII
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
  B.2.&.
  Continued
  E.l.Za
     The temperature of the samples and standards
must be the same before carrying out the test.
Consequently if a difference in temperature exists,
allow both to sit at room temperature for about one
hour before proceeding with the test.

     Analytical  results obtained with the SPADNS
reagent are limited to a range of 0 to 1.4 mg F"/
liter.  Samples  which approach or exceed the limit of
the range must be diluted.  The dilution must be
made before addition of reagent, since subsequent
dilution will affect the concentration of dye,
zirconium and acid.

     The first method to calibrate the spectrophoto-
meter is to prepare a calibration graph.  Calibration
graphs are commonly used in absorbance measurements.
In this type measurement energy is absorbed by some
chemical constituent in the solution by means of a
calibration graph.

     Two things  must be done in order to prepare a
calibration graph.   A series of standards must be
prepared.  A standard is a solution which contains
a known amount of the same chemical constituent
which is being determined in the sample.  Secondly,
the absorbance of these standards must be measured.

     This is done by carrying the standards through
the test procedure  and measuring the amount of energy
absorbed.  This  value is graphed against the known
concentration and a line drawn through the points
produced.  This  line is the calibration graph.
When an unknown  sample is run, its absorbance value
is determined and using the calibration graph, its
concentration can be determined.

     The second  method used in the SPADNS uses the
fact that a straight line will be formed between 0
and 1.4 mg F/liter.  Then the equation for a straight
line is used to  calculate the unknown concentration
after values for two standards have been run.

     This is carried out as follows.  Two standards
are prepared, usually 0 and 1 ppm.  These are used to
set the instrument, thus obtaining values for their
absorption.  These  two knowns can be used to
calculate the unknown by the equation.  Both methods
are shown in the EMP.
                                                                                Ell.A-21

-------
EFFLUENT MONITORING PROCEDURE;
                Determination
                SPADNS Method
of Fluoride Using the
  FIELD AND LABORATORY ANALYSIS
                                                                           Section
                                  TRAINING  GUIDE  NOTE
                                                       REFERENCES/RESOURCES
  E.l.Bb
  E.Z.lOa
     The addition of the highly colored SPADNS
reagent must be done with utmost accuracy because
the fluoride concentration is measured as a
difference of absorbance in the standard and the
sample.  A snail error in reagent addition is the
most prominent source of error in this test after
the interferences.

     The absorbance scale is logarithmic.  This
means that the distance between numbers varies.  The
meter distance between numbers is larqe on the right
side of the meter while the distance between numbers
is small on the left side.  For example the distance
between 0 and 0.5 is over half the scale while from
1  to 2 is less than a quarter of the total distance.

     Consequently the right side of the scale can
be read with a much hinher degree of sensitivity and
accuracy.

     The ratio of fluoride concentration and absorb-
ance is inverse, that is the higher the fluoride
concentration the lower the reading and the lower
the fluoride concentration the higher the reading,
Thus by setting the lowest concentration of fluoride
possible, i.e., 0 mg F~/1iter, at 0.5 absorbance
units this means that no reading can go higher.
This restricts all  readings to the most sensitive
section of the scale.
                        Methods for Chemical
                        Analysis of Water and
                        Wastes 1974, E.P.A.
                        Environmental  Monitoring
                        and Support Laboratory
                        Cincinnati, OH 45268.
     Ell.A-22

-------
TYPICAL LABORATORY DATA SHEET                             Section IX

Sample Collection:
1.  Name of Plant:___	,	
2.  Sampling Location:	
3.  Type of Sample:__	„	,	
4.  Date and Time Collected:
5.  Sample Collector:	
SampleAnalysis:
1.  Name of Laboratory:
?..  Date and Time Collected'
3,  Sample Designation:	
4.  Method Used:	
5.  Was Sample Distilled:	
6.  Concentration of F~ Determined (in ing/liter):
                                                                     Ell.A-23

-------
                          A  PROTOTYPE  FOR  DEVELOPMENT  OF
                          ROUTINE  OPERATIONAL  PROCEDURES

                                     for the

               DETERMINATION OF  FLUORIDE IN POTABLE  AND WASTEWATERS
                         USING A SELECTIVE ION ELECTRODE
                                as applied  in

                          WATER TREATMENT  FACILITIES

                      WASTE  WATER TREATMENT  FACILITIES

                                 and  in the

                    MONITORING OF EFFLUENT  WASTEWATERS
              National Training and Operational Technology Center
                       Office of Water  Program Operations
                      U.S.  Environmental  Protection Agency
CH.HAL.f.lab.WMP.2.11.77
                                                                           Ell.B-1

-------
EFFLUENT MONITORING PROCEDURE:   Determination  of Fluoride  in  Potable  and
                                Waste Waters Using  a  Selective  Ion  Electrode

This operational  procedure was  developed by:

     NAME        John D.  Pfaff

     ADDRESS     USEPA,  OWPO, National  Training  and Operational  Technology  Center
                 Cincinnati, Ohio  45268

     POSITION    Chemist - Instructor

     EDUCATION AND TECHNICAL BACKGROUND

                 B.A.  Chemistry

                 3 years  Research  Chemist

                 14 years DHEW, ECA,  EPA -  Chemist
                                                                     Ell.a-3

-------
        EFFLUENT MONITORING PROCEDURE:   Determination  of  Fluoride  in  Potable
                                        and  Waste  Waters  Using  a Selective  Ion  Electrode

        1.  Analysis  Objectives:

           The learner will  determine  the  fluoride content  of a water or wastewater
           samole using a  selective ion  (fluoride) electrode.

        2.  Brief Description of  Analysis:

           The nH of the samole  is  adjusted  to  between 5  and 5.5 by use of
           buffer solution.   Then the  electrodes are  immersed in the  solution and
           the fluoride content  determined using the meter.  If an expanded scale
           pH meter  is used the  concentration  is read  from  a standard curve.  If a
           specific  ion meter is used  the  concentration can be  read directly from the
           face of the meter. Other acceotable methods not covered in this outline
           are the SPADNS  and automated  complexone procedures.

        3.  ADD!icability of this procedure:

           a. Range  of Concentration:

              from Q.I to  1000 mg/1
              Information  is qiven  so  the  same  stepwise procedure  can be used for
              fluoride concentrations  UD to  1000 mg/liter.

           b. Pretreatment of Samples:

              Distillation is not necessary  for potable water samples.  However,
              the guidelines for MPDES procedures  soecify distillation of waste
              water  samples must be carried  out.   (See training quide note  VII .B.I.la.).
              The distillation procedure is  not included  but is covered in  another EMP
              on Fluoride  Distillation.

           c. Treatment of Interferences in  Samples:

              Interferences are  few when the buffer is used with the  electrode,

              The polyvalent cations of  si ",  fe   , al  , interfere but can be
              tolerated up to 5,0 mq/1 with  use of the buffer.  Extremes of pH
              can cause problems but adjustment of pH  with  the  buffer negates
              this problem.

              Additional information is  given  in the  training guide note VII.E.4.
        *Source of Procedure:   Methods  of Chemical  Analysis  of  Water  and  Wastes,  1974,
        Environmental  Protection Agency,  Environmental  Monitoring  and Support  Laboratory,
        Cincinnati, Ohio 45268, p.65.
Ell.B-4

-------
EFFLUENT MONITORING PROCEDURE:   Determination of Fluoride in  Potable  and
                                Waste Waters Using a Selective Ion  Electrode

Equipment and Supply Requirements

A. Capital Equipment:

   1.  Expanded scale pH meter or selective ion meter

      Examples:  Beckman: Expandomatic Model  76007
                Coleman: Model  37A
                Corning: Model  12
                Fisher Accumet Model  320
                Leeds & Northrup Model  7405  - A 2.
                Orion Model  407A portable
                Hach pH/Fluoride meter No. 12330
                                       No.12320 - portable

   2.  Sleeve-type reference  electrode

      Examples:  Beckman: No.  40463
                Coleman: No.  3-721
                Corning: Mo.  476012
                Fisher:   No.  13-639-62
                Orion:   No.  90-01

   3.  Fluoride Electrode

      Example:   Beckman:  39600
                Coleman:  3-803
                Corning:  476042
                Orion:    94-09
                Orion:    96-09
                Hach:     13034-00

   4.  Trip balance,  500  gram  capacity.

   5.  Magnetic stirer  and teflon covered  stir  bar, about  2.5 cm long.

   6.  Water still or other source of  distilled water.
                                                                     Ell .B-5

-------
        EFFLUENT MONITORING PROCEDURE:
        B,  Reusable Supplies:
Determination of Fluoride in Potable and
Waste Waters Using a Selective Ion Electrode
           1. One  stop watch, clock or watch  (with second hand).
           2. One  thermometer, glass 0 to 100 C.
           3. One  elastic squeeze bottle
           4. One  stirring rod, glass about in inches lonq.
           5. One  pair, safety glasses
           6. One  powder funnel, glass about  3 inch diameter
           7. One  laboratory aoron
           8. Five weighing boats, plastic  (2-3 inches square),
           9. One  note book (for recording data)
          10. One  pen or pencil
          11. One  Flask volumetric 1000 ml volume.
          12. Two  Flask volumetric 100 ml volume
          13. Ten  Flasks 50 ml volume (for use with pH meter only)
          14. One  Cylinder graduated 500 ml volume
          15. One  Cylinder graduated TOO ml volume
          16. Seven  Pipet volumetric 10 ml
          17. Two  Pipet graduated 10 ml
          18. Four Beakers, plastic 100 ml volume
          19. One  Pipet Bulb
          20. One  spatula

        The following will be needed in addition to the above only if the buffer is
        prepared rather than purchased.

          21, One  Erlenmeyer flask, 500 ml volume
          22. One  Beaker, 1000 ml volume
          23. One  Flask volumetric, 1000 ml volume
          24. One  pH electrode

        C.  Consumable Supplies:

           1. a. Sodium Fluoride -NaF Reagent grade powder - 4  oz.or
             b. Sodium Fluoride Stock Solution -
                Orion Research Inc. 380 Putnam Ave. Cambridge, Mass 02139, Cat. Mo.
                94-06-07
                Hach Chemical Co, P.O. Box 907 Ames Iowa, 50010, Cat. No. 232-11

           2. Adjustment buffer

             a. Total Ionic Strength Adjustment Buffer (TISAB)
                Orion Research Inc. 380 Putnam Ave, Cambridge, Mass.  02139 Cat. No.
                94-09-09

             b. Fluoride Adjustment Buffer,  Formula - 2589
                Hach Chemical Co. P.O. Box 907 Ames Iowa,  50010 Powder - Cat. No.
                                                                          2589.01
                                                                 Pillows - Cat. No.
                                                                          2509-99
Ell.B-6

-------
EFFLUENT MONITORING PROCEDURE:   Determination of Fluoride in  Potable  and
                                Waste Waters Using a Selective Ion  Electrode

C. Consumable Supplies (Continued):

   3. The following are needed  if the adjustment buffer is prepared instead
      of purchased,

      a. Acetic Acid, Glacial,  CH,COOH.   Reaqent qrade - 1 pt
      b. CDTA* (1,2.  cyclohexylene dinitrilotetraacetic acid) 25  q.
               Matheson,  Coleman A  Bell  Cat. No.  CX - 2390
      c. Sodium Chloride, NaCl,  Reagent  grade - 1  oound
      d. Sodium Hydroxide, NaOH, Reagent grade - 1 oound
* Also listed as 1,2~cyclohexylene" diaminetetraacetic  acid Baker  Cat. No. G083,
                                                                    Ell.B-7

-------
                                              FLUORIDE ANALYSIS FLOW SHEET
                                                                          E11.B-8
                                                                                Sample
                                                                   Potable  Water
                                                         Waste Water
                                                                                                  Distill
Electrode

   Method
       SPADMS

Colorimetric  Method
    SPADf.'S

 Colorimetric
	Method
Automated

Complexone
Method

-------
EFFLUENT MONITORING PROCEDURE.:
                                Determination of Fluoride in Potable and  Waste Water
                                                  Ion Electrode
 OPERATING PROCEDURES
                                  STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Sample Collection
B. Sample pretreatment
C. Equipment Preparation

   1, Glassware
   2. Electrometer
      (pH meter or
      selective ion meter
                          1. Collect a minimum of 300
                             ml in a plastic or hard
                             glass container.
la.  Polyethylene bottles are preferred
    Glass bottles are satisfactory provided they have
    not previously contained high-fluoride solutions.
Ib.  No special  requirements are necessary for
    preservation.
Ic.  Chlorine does not interfere so no precautions are
    necessary.
                          1, Distill sample
la. Distill  wastewater sample for total  or total
    dissolved fluoride.
Ib. Use the  EMP covering the distillation procedure
    for Fluoride.
Ic. Distillation is not required for drinking water
    samples.
                           1.  Clean all glassware and
                              plastic  beakers  in
                              detergent.

                           2.  Rinse with distilled water

                           1.  Check meter  zero.

                           2.  Check power  supply.
                           3. Check connection of
                              electrodes to meter,
la. Most instruments have a mechanical  screw
    adjustment to center the pointer of the meter face
2a. Check correct connection to voltage source for
    line operated meters.  Check batteries on all
    portable and some line operated models.  See
    instrument manual for directions.
3a. There should be two connecting pins that attach
    to the meter.  The longer one from the measuring
    electrode, is the fluoride electrode, and the
    smaller one is from the reference electrode.
                                                                                                              Ell.B-9

-------
 EFFLUENT MONITORING PROCEDURE
      Determination of Fluoride in Potable and
      Waste Waters Using a Selective Ion Electrode
                                                                                                           Ell.B-10
 OPERATING PROCEDURES
        STEP SEQUENCE
                  INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE NOTES
3. Reference Electrode
D.  Reagent Preparation

   1.  Distilled Water
      Sodium Hydroxide
      NaOH,  5N
      (for use in
      preparation  of
      buffer)
   3.  Buffer  Solution
      (preparation
      optional  -  can
      he  purchased)
1. Check Internal
   solution.
fillinq
3b. Some connections are spring loaded and must be
    held in the socket until clamped with a screw.
3c, Combination electrodes are two electrodes placed
    one inside the other.  This electrode will  still
    have two connecting pins.

la. Before usinq the reference electrode make sure the
    internal filling solution has been added.  This
    solution is provided with the electrode and must
    be used.
Ib. Fill about half the electrode with the solution.
   Prepare about six (6)
   liters of distilled water.
   This water must be free
   from fluoride.
1. Weigh out 40 grams of
   sodium hydroxide (NaOH)

2. Dissolve the sodium
   hydroxide in 200 ml  of
   water in a 500 ml
   Erlenmeyer flask.

1. Add approximately 590 ml
   of distilled water to a
   1  liter beaker.
             la.  Use a still  or pass tap water through an ion-
                 exchange column containing  a  strongly acidic
                 cation exchange resin mixed with a strongly
                 basic anion  exchange resin.
             la.  Use a  trip balance.
             Ib.  Put sodium hydroxide in a plastic  weighing boat
                 for weighing.
             2a.  Use a  graduated cylinder to measure the  200 ml of
                 water.
             2b.  Caution:  heat  given  off.
             la.  This  solution can be purchased  already prepared.
                 Sources  are listed in the front of the EMP.

-------
EFFLUENT MONITORING PROCEDURE:
       Determination  of  Fluoride  in  Potable  and
       Waste  Waters Using  a  Selective  Ion  Electrode
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Reagent Preparation

   3.  Buffer Solution
      (Continued)
2.  Add 57 ml  of glacial
   acetic acid.
                          3. Weigh out 58 grams of
                             sodium chloride (Nad)
                          4. Add sodium chloride
                             (NaCl) to the 1  liter
                             beaker,

                          5, Weigh out 4 grams of CDTA.
                          6. Add CDTA to the 1 liter
                             beaker.

                          7. Stir to dissolve.

                          8. Place beaker  in a cool
                             water bath.

                          9. Insert calibrated pH
                             electrode  into solution.
2a, Use a graduated cylinder.
2b. Caution:  Use in a well  ventilated area.
2c. Use safety glasses.
                                3a.  Use a trip balance.
                                3b.  Put sodium chloride  in a plastic weighing  boat
                                    for weighing,
                                3c.  For work with  brines additional  Nad  should be
                                    added to raise the chloride level to  twice the
                                    highest expected level of chloride in the  sample,

                                4a.  That is the same beaker to which the  glacial
                                    acetic acid has been added.
                                5a. The E.P.A. Methods Manual  1974, calls for 2 grams
                                    however, STD Methods 14th and the Orion manual
                                    both call for 4 grams.
                                5b. Use trip balance-
                                5c. Put CDTA in a plastic weighing boat for weighing.
                                5d. CDTA - 1,2 cyclohexylene diamine tetraacetic
                                    acid.
                                8a. Cool to room temperature.
                                9a. The sane meter that will be used for the fluoride
                                    measurement can be used for this.

                                9b. Consult EMP on pH measurement.
                                                                                                             E11.B-11

-------
EFFLUENT MONITORING PROCEDURE:
       Determination  of  Fluoride  in  Potable  and
       Waste Waters Usinn  a  Selective  Ion  Electrode
                                                                                                         Ell.B-12
OPERATING PROCEDURES
         STEP  SEQUENCE
     INFORMATION/OPERATING  COALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Reagent Preparation

   3.  Buffer Solution
      (Continued)
   4.  Sodium Fluoride
      (NaF)  stock
      solution 100 mg
      F"/liter
10.  Add  5N Sodium hydroxide
    (Reagent D2)  (NaOH)  to  the
    solution until  meter reads
    5.0  to 5.5 pH.
                         11.  Allow solution  to  reach
                             room temperature.

                         12.  Transfer solution  to a  1
                             liter volumetric  flask.
13.  Dilute to volume and mix
    thoroughly.

 1.  Weigh out 0.2210 grams of
    sodium fluoride.
                          2.  Add about 500 ml  of
                             distilled water to a 1
                             liter volumetric  flask.

                          3.  Transfer the solid to the
                             1  liter volumetric flask
                             using a powder funnel.

                          4.  Use wash bottle to wash
                             the solid into flask.

                          5.  Dissolve the solid.

                          6.  Dilute to volume  and mix
                             thoroughly.
lOa.  Reagent 2.
lOb.  About 150ml  will  be needed.
ICc.  Use a 100 ml  graduated cylinder to add the sodium
     hydroxide until  the pH nears  5.0.   Then add from
     a pipet dropwise until the range of 5.0 to 5.5
     is reached.
                                12a.  Use wash bottle and  rinse  beaker  with  small
                                     amounts  of distilled water,  transferring  this
                                     wash water to  the volumetric  flask.
 la. Use analytical balance.
 Ib. Use plastic weighing boat.
 Ic. Solution can be purchased from Orion Research
     Inc., Cat NO. 94-06-07 or Hach Chemical Co., Cat.
     No. 232-11

 2a. Approximately \ full.
                                 4a.  The weiging boat should be washed three times
                                     and the washings added to the flask.
                                 6a.  Solution contains 0.1  mg F~ per 1.0 ml
                                     (i.e.  IOC ppm F")

-------
EFFLUENT MONITORING PROCEDURE!:    Determination  of  Fluoride in Potable and
                           '~    Waste Waters  Using  a  Selective Ion Electrode
OPERATING PROCEDURES
        STEP SEQUENCE
INFORMATION/OPERATING  COALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Reagent Preparation

   4. Sodium Fluoride
      (NaF) stock
      solution 100 mg
      F~/liter
      (Continued)
   5. Sodium Fluoride
      (NaF) Solution
      10 mg F~/liter
    6.  Sodium  Fluoride
       (NaF) solution
       1.0  mg  F"/liter
 E. Meter and Electrode
    Check
7.  Label  container.
1. Add approximately 50 ml
   distilled water to 100 ml
   volumetric flask.

2. Pipet 10 ml of the (NaF)
   stock solution (reagent
   C.4.) into the flask.

3. Dilute with distilled water
   to the mark.

4. Mix  thoroughly.

1. Add  approximately 50 ml  of
   distilled  water  to a 100
   ml volumetric  flask.

2. Pipet 10 ml of reagent  5
   (10  mg  F~/l)  1iter  into
   the  flask.

3. Dilute  with distilled
   water to  the  mark.

4. Mix  Thoroughly.
                                6b.  Keep in plastic.   Stable for 6 months.
                                                          2a. Use a 10 ml volumetric pipet.
                                                          3a. Solution contains 0.01 mg F~ Der  if) m] i.e.,in
                                                              ppm.  Label solution.
                                                           2a.  Use  a  10  ml  volumetric  pipet.
                                                           3a.  Solution  contains  0.001  mg  F~  per 1.0  ml  i.e.,
                                                               1.0 ppm.

                                                           4a.  Label
 1.  Turn on Meter
                                                                                                              Ell .B-13

-------
EFFLUENT MONITORING PROCEDURE:
       Determination of Fluoride in Potable and
       Waste Waters Using a Selective Ion Electrode
                                                                                                          Ell.B-14
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE  NOTES
E.  Meter and Electrode
   Check  (Continued)
2. Allow to warm up.

3. Pipet 10 ml  of Reagent 6
   into a 100 ml plastic
   beaker.

4. Add 10 ml  of buffer i.e.,
   Reagent 3.
2a.  If battery powered, no warm up necessary.

3a.  Use a 10 ml  volumetric pipet.
                                                          4a.
                                                          4b.
                                                          4c.
                                                          4d.
                          5.  Place  a  stirring  bar  into
                             the  beaker.

                          6.  Place  beaker  on stir  plate

                          7.  Lower  electrode or
                             electrodes  into solution.
                                7a,
                                                          7b.
                          8.  Turn  on  stir  plate.
                                8a.
                                                          8b.
                          9. Turn meter on  to a
                            millivolt reading position
    Use a 10 ml  volumetric pipet.
    Any volume of sample or standard and buffer can
    be used, provided equal volumes are used.   For
    example 10 ml of sample plus 10 ml  of buffer.
    Use of powdered buffer (Hach Co.)  eliminates the
    dilution of sample or standard  and  therefore
    eliminates possible error.
    Volume should be sufficient to  cover the electrode
    or electrodes to a depth of about  1  inch.
VII.E.4
VII.E.4b
(pp.  27
& 28)
    Care should be taken that the stirring bar does
    not hit the electrode.
    Before using the reference electrode make sure
    the internal filling solution has been added.
    This solution is provided with the electrode and
    must be used.  Fill  about half the electrode with
    the solution.

    For best results stirring should be at a rate
    that will  not cause  a vortex.
    Insulate with cardboard or styrofoam between
    sample and stir plate to reduce possibility of
    sample temperature change.  Not so thick as to
    stop the stirrer.
                                9a.  The  meter should  be used  in  the expanded  scale
                                    mode.
V.E.7.
(p. 26)

-------
EFFLUENT MONITORING PROCEDURE:
        Dctonnination  of Fluoride in  Potable  and
        Waste Waters  Using  a  Selective  Ion  Electrode
OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E.  Meter and Electrode
   Check  (Continued)
10.  Record reading in
    mil 11 volts.
                         11.  Turn meter to standby
                             position or off position,

                         12.  Raise electrodes from
                             solution.

                         13.  Rinse electrodes with
                             disti11ed water.

                         14.  Blot dry with soft tissue.

                         15.  Pipet 10 ml of Reagent 5
                             into a clean 100 nil  plastic
                             beaker.

                         16.  Add 10 ml buffer.
                         17. Place a stirring bar into
                             the beaker.

                         18. Place beaker on stir
                             plate.

                         19. Lower electrodes into
                             solution.

                         20. Turn meter on at a
                             millivolt reading position
lOa.  The electrodes must remain in the solution for at
     least three minutes or until  the reading has
     stabilized.  At concentrations under 0.5 mg
     F /liter it may require as long as five minutes
     to reach a stable meter reading.  Higher
     concentrations stabilize more quickly.

lla.  Depends on type meter being used.
                                15a. Use a 10 ml volumetric pipet.
                                16a.
                                16b.
                                19a,
     Use a 10 ml volumetric pipet.
     Samples and standards should be at the same
     temperature.  A 1°C difference in temperature
     will give rise to about a 2 X error.
     Care should be taken that the stir bar does not
     hit the electrode.
                                                                                                             Ell.6-15

-------
EFFLUENT MONITORING PROCEDURE
Determination of fluoride in Potable and
Waste Waters Using a Selective Ion Electrode
                                                                                                        Ell.B.16
OPERATING PROCEDURES
E. Meter and Electrode
Check (Continued)

F. Calibration
1 . Using a pH meter




STEP SEQUENCE
21. After 3 minutes record the
reading in mil 1 ivol ts.
22. Determine the difference
between the first and
second millivolt reading.

1. Turn on meter and allow to
warm up.
2. Prepare a series of
standards using Reagent 5.


3. Pipet 10 ml of a standard
into a 100 ml plastic
INFORMATION/OPERATING COALS/SPECIFICATIONS
21a. Or until the meter
stabil izes .
22a. Correct electrode operation is indicated by a
difference of about 58 millivolts, assuming the
solution temperature is between 20°C and 25 C.
22b. If the change is not within + 2 millivolts,
consult the electrode manual.




2a. The following table can be used to prepare a
series containing from 0 to 2.0 mg F /liter by
diluting appropriate volumes to 50 ml.
ml of Reagent 5
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
10.00
Cone, when diluted to
50 ml in mg F~/l iter
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
1.60
2.00
2b. Use a 10 ml graduated pipet.
3a. Any volume can be used provided equal amounts of
sample or standard and buffer are used.
TRAINING
GUIDE NOTES

III. E. 22
(p. 25)






                             beaker.

-------
EFFLUENT MONITOR ING PROCEDURE:    Determination  of  Fluoride  in  Potable  and
    "  	                         Waste  Waters  Using  a  Selective  Ion  Electrode
OPERATING PROCEDURES
         STEP  SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F,  Calibration

   1.  Using a pH meter
      (Continued)
   2.  Using a selective
       Ion Meter
                          4.  Add 10 ml  of buffer.

                          5.  Add a stir bar to  the
                             beaker.

                          6.  Lower the  electrodes  into
                             the beaker.

                          7.  Turn on stir plate.
 8.  Turn on  meter.


 9,  Allow meter to  stabilize.

10.  Record millivolt  reading.

11.  Turn meter to off or
    standby  position,

12.  Raise electrodes.

13.  Rinse the electrodes.

14.  Blot dry.

15.  Repeat steps 3-13 for  each
    standard,

16.  Prepare  a standard curve.

 1.  Pipet 10 ml of  Reagent 6
    into a 100 ml plastic
    beaker.
                                3b.  Use a  10 ml volumetric pipet.

                                4a.  Use a  10 ml volumetric pipet.
 7a.  Do  not  allow  stir  bar  to  hit  electrode.
 7b.  Stir at a  rate  that will  not  cause  a  vortex.

 8a.  Use expanded  mode.
 8b.  Use millivolt mode.

 9a.  About 3 minutes.
                                                         lla.  Always go to either position,  depending  on  meter,
                                                              before raising the electrodes  from the  solution.
13a.  With distilled  water.

Ha.  With soft tissue.

15a.  Repeat for each standard until  all  are run,
     recording the millivolt reading.

16a.  Plot on two cycle semilog paper.

 la.  Use a 10 ml volumetric pipet.
                                                                                                                 VII  F.l.lSa
                                                                                                             Ell.B-17

-------
EFFLUENT MONITORING PROCEDURE:
Determination of Fluoride in Potable and
Waste Waters Using a Selective Ion Electrode
                                                                                                         m.B-18
OPERATING PROCEDURES
F.  Calibration

   2.  Using  a selective
      Ion Meter
      (Continued)
 STEP SEQUENCE
                          2.  Pipet  10 ml  of  buffer  into
                             the beaker.

                          3.  Place  a  stirring  bar  into
                             the beaker.

                          4.  Place  beaker on stir  plate.

                          5,  Lower  electrodes  into
                             beaker.

                          6.  Turn on  stir plate.

                          7.  Turn meter  to the mono-
                             valent anion position.

                          8.  Using  the calibration  con-
                             trol adjust the meter  to
                             read at  center  scale.

                          9.  Turn meter  to off.

                         10.  Raise  the electrodes.

                         11.  Rinse  electrodes  with
                             distil led water.

                         12.  Blot dry with soft tissue.

                         13.  Pipet  10 ml  of  Reagent 5
                             into a clean plastic  100 ml
                             beaker.
INFORMATION/OPERATING GOALS/SPECIFICATIONS
                         lb.  Any volume can be used provided equal  amounts of
                             sample or standard and buffer are used.

                         2a.  Use a 10 ml  volumetric pipet.
                         5a. Do not allow stir bar to hit electrode.
                         6a. Stir at a rate that will  not cause vortex.
                         7a.
                         7b.

                         8a.
                         8b.
This may be marked as X  or F .
Allow three minutes before reading.

Use the logrithmic scale on the meter.
This adjustment makes the_instrument read a
concentration of 1.0 mg F~/liter at center scale.
                        13a.
                        13b.
Use a 10 ml volumetric pipet.
This solution will be a 10 mg F~/liter standard.
  TRAINING
 GUIDE  NOTES
VII.F.2.8a.
(p. 29)

-------
EFFLUENT MONITORING PROCEDURF:
       Determination  of  Fluoride  in  Potable  and
       Waste  Waters Using  a  Selective  Ion  Electrode
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING COALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Calibration

   2. Using a selective
      Ion Meter
      (Continued)
14.  Add 10 ml  of buffer.

15.  Place stir bar into
    beaker.

16.  Place beaker onto stir
    plate.

17.  Lower electrodes into
    solution.

18.  Turn meter to monovalent
    anion position.

19.  Allow to stir for three
    minutes.

20.  Use Temperature
    compensator to adjust
    meter needle to the last
    number on far right of log
    scale.
14a.  Use a 10 ml  volumetric pipet.
                                                         17a. Do not allow stir bar to hit electrodes.
                                                         19a. Or until meter stabilizes.
                                                         20a. The temperature of all samples and standards
                                                              should be the same.  If sample is not the same
                                                              temperature, allow to stand at room temperature
                                                              for about one hour.

                                                         20b. This adjustment makes the meter read a con-
                                                              centration of 10.0 mg F~/liter at far right
                                                              scale.
                                                         20c. The entire scale now can be used to read directly
                                                              concentrations from 0.1 rig F /liter to
                                                              10.0 mg F~/liter.
                                                         20d. This calibration should be checked about every
                                                              1  to 2 hours by repeating procedure F2.  If the
                                                              temperature changes, recalibrate more often.
                                                         20e. The log scale may  be  located at various positions
                                                              on the meter face.  Consult the manufacturers
                                                              manual to point out which to use if in doubt.
                                                                                                             Ell.B-19

-------
EFFLUENT MUKITORING PROCEDURE:
        Determination of Fluoride in Potable and
        Waste Waters Using a Selective Ion Electrode
                                                                                                          Ell.B-20
OPERATING PROCEDURES
         STEP SEQUENCE
INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F. Calibration

   2. Using a selective
      Ion Meter
      (Continued)
G. Procedure
 21,  Turn meter off.

 22.  Raise electrodes.

 23.  Rinse electrodes with
     distil 1ed water.

 24.  Blot dry with soft tissue
 1. After the calibration of
    the meter has been com-
    pleted, test all  samples
    by doing the following.

 2. Turn on meter for warm up.

 3. Pipet 10 ml  of sample into
    a clean plastic 100 ml
    beaker.

 4. Add 10 ml  of buffer.

 5. Place stir bar into beaker.

 6. Place beaker on stir plate.

 7. Lower electrode.

 8. Turn on stir plate.

 9. Adjust stir  plate to a
    rate that will not form
    vortex.

10. Turn on meter.
                                                          2a.  If necessary.

                                                          3a.  Any volume can be used  provided  equal  amounts  of
                                                              sample or standard and  buffer are  used.
                                                          3b.  Use a  10 ml  volumetric  pipet,

                                                          4a.  Or same volume as used  for  sample.
                                                          8a.  Do not allow stir bar to  hit electrode.
                                                         lOa.  To monovalent am'on  position.

-------
EFFLUENT MONITORING PROCEDURE:
       Determination of Fluoride in Potable and
       Waste Waters Using a Selective Ion Electrode
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                    TRAINING
                                                   GUIDE NOTES
G.  Procedure
   (Continued)
1.  Stir for three minutes,  or
   until  meter stabilizes.

2.  Read value.
H. Storage

   1. Reference
      Electrode
    2.  Fluoride?
       Electrode
                                                         12a.  For  pH  meters  read  at  millivolt  scale  and  use
                                                              millivolt  vs  concentration  curve  to  determine
                                                              F~  concentration  in sample.
                                                              For  selective  ion meter read  concentration
                               12b.
                         13.  Turn meter off.

                         14.  Raise electrodes.

                         15.  Rinse and dry electrodes.
   Store in the 1.C mg F /
   1iter standardizing
   solution or distilled water
   if the electrode is to be
   used in a short time or
   next day.

   Clean thoroughly and store
   dry if storing for long
   period of time.

   Store in air  or in a
   standardizing solution.
 I.  Calculations

    1.  pH  meter
 1.  Use  the  constructed
    standard curve.
                                                              directly from logarithmic scale.
la.  If using separate electrodes,  i.e.  reference
    and fluoride,  this section  applies  to  the
    reference only.   If a combination electrode  is
    being used the precautions  for the  reference
    electrode section should be used.
                                                          2a. Before reuse add internal filling solution.
la. No specific storage precautions necessary.
la.
When the reading of millivolts is obtained
this value to convert to concentration in
      T -; -*- ~ ~
                                                                      obtained use
                                                              mg F" liter.
                                                                                                              Ell.B-21

-------
EFFLUENT MONITORING PROCEDURE:
    Determination of Fluoride in Potable and
    Waste Waters Using a Selective Ion Electrode
                                                                                                          Ell.B-22
OPERATING PROCEDURES
     STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
I.  Calculations

   1.  pH meter
      (Continued)
   2,  Specific Ion
      Meters
Read concentration scale
(logarithmic directly to
obtain sample mg F /liter.
Ib.  Using the graph paper attached.   Look at the
    bottom of the graph and find the millivolt
    reading for the sample.  Move upward until  the
    curve is contacted then move horizontally left
    until the edge of the paper is reached.   Read
    this intersection in mg F~/ liter.
Ic.  If additional instruction on the use of  a
    calibration graph is needed, consult the EMP,
    Preparation of Calibration Graphs.

la.  All data reported in mg F"/liter.

-------
                                TRAINING GUIDE

SECTION                         TOPIC
   I*                Introduction
   II                Educational Concepts - Mathematics
   III*              Educational Concepts - Science
   IV                Educational Concepts - Communications
   V*                Field & Laboratory Equipment
   VI                Field & Laboratory Reagents
   VII*              Field & Laboratory Analysis
   VIII              Safety
   IX*               Records & Reports
*Training guide materials are presented here under the headings marked*.
These standardized headings are used through this series of procedures.
                                                                           Ell.B-23

-------
EFFLUENT MONITORING PROCEDURE:
                Determination of Fluoride in Potable and
                Waste Waters  Using a Selective Ion Electrode
INTRODUCTION
                                                                           Section
                                  TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
                The basis  for  this
                electrode  itself,
                fluoride solution;
                crystal  doped  with
                   method is in the fluoride
                   Most electrodes contain a
                   at the tip of the electrode is a
                   fluoride ions.   The crystal acts
as an ion-exchange membrane, so that when the
fluoride concentration outside of the electrode is
higher than that inside, ions move toward the inside
setting up a voltage potential proportional  to the
difference in fluoride concentration.  Conversaly
when the fluoride concentration on the outside is
lower than that on the inside, a proportional
potential of opposite sign is set up.  In most
fluoride electrodes, the internal  solution is about
10   molar in fluoride, so concentrations below
19 mg F~/liter result in positive voltage readings.
Some electrodes contain no internal solution, but the
principle of operation is similar.  However, in the
solid electrodes the potential developed by a
particular fluoride solution is independent of a
filling solution but rather depends entirely on the
characteristics of the particular crystal used in the
electrode manufacture.

-------
EFFLUENT MONITORING PROCEDURE:
               Determination of Fluoride in Potable and
               Waste Waters Using a Selective Ion Electrode
 Educational  Concepts - Science
                                                                           Section
                                                                   III
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 1.22
Theory of operation of a fluoride electrode predicts
a behavior that is expressed by the Nernst equation.
Consequently the difference in millivolt readings
between two concentrations of fluoride that differ
by a factor of ten should be between 55 and 60
mil 1ivolts,

If one knows concentration of fluoride, for example
1 mg/liter, is expressed on a meter in millivolts anc
then a second concentration differing by a factor of
ten, for example 10 mg/liter, is expressed on the
same meter in millivolts, the difference between
these millivolt values will be about 58 millivolts.

This fact can be used to check on whether the system
is operating properly.  How this is carried out will
depend on the type of meter and electrode being used.
                                                                                   Ell.B-25

-------
EFFLUENT MONITORING PROCEDURE:   Determination  of  Fluoride in  Potable  and
	   Waste Waters  Using  a  Selective Ion  Electrode
Field & Laboratory Equipment
                                                                           Section
                                  TRAINING GUIDE  NOTE
                                                       REFERENCES/RESOURCES
E.7
The fluoride electrode must be used in conjunction
with a standard single junction sleeve-type
reference electrode.  If smaller volumes of sample
are to be run for convenience sake, a combination
electrode is available.  This appears to be a single
electrode.  However, in actuality, it is two
electrodes, one inside the other.  This allows the
electrode to be inserted into smaller diameter
containers and also makes for easier use.
     Ml

-------
EFFLUENT MONITORING  PROCEDURE:
               Determination of Fluoride in Potable and
               Waste Waters Using a Selective Ion Electrode
Field & Laboratory Analysis
                                                                           Section
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 1.1,la,
 E, 4.
For potable waters to comply with requirements as
listed in the National Interim Primary Drinking
Water Regulations (F.R. Part IV - 12/24/75-§10 - pg
59573), distillation is not necessary.

However, for surface and saline waters or domestic
and industrial wastes to comply with NPDES
requirements distillation if necessary unless
sufficient data exists to prove that distillation is
not necessary.  Manual distillation will  be required
to resolve any controversy.

Interferences connected with the electrode can be
1 is ted as being:

a) PH
b) temperature
c) other ions
d) total ionic strength

In acid solutions below pH 5, hydrogen can complex
fluoride by forming the undissociated acid
hydrofluoric HF.  This will not allow the electrode
to sense the fluoride concentration, thus tied up,
Hydroxide ion can also interfere when concentration
of this ion are about 10"^ Molar.  Addition of the
buffer solution will adjust most samples  into the
range of 5.0 to 5.5 pH where no interference of pH
is found.  For samples high in pH, i.e.,  pH 11, the
volume of buffer added may not be sufficient.
These samples should have their pH adjusted by the
addition of IN HC1 to adjust the pH to about 8 before
addition of the buffer.

So long as the temperature of the standards and
samples are the same, temperature will not affect the
readings.  However a variance between standard and
sample of 1 C can cause a 2% error for a  concentra-
tion of 19 mg F/l.

Other ions, particularly, iron and aluminium can
cause significant errors.  The table below gives
some indication of this error.  The numbers given
show the concentration that will affect the
electrode after addition of buffer.
                                                                                  Ell.B-27

-------
EFFLUENT MONITORING PROCEDURE:
               Determination of Fluoride in Potable and
               Waste Waters Using a Selective Ion Electrode
Field & Laboratory Analysis
                                                           Section
                                  TRAINING GUIDE  NOTE
                                                       REFERENCES/RESOURCES
 E.4
.Continued)
E.4.b.
F.1.15a.
       INTERFERENCES FOR THE ELECTRODE

Concentration of substance, in mg/liter required to
cause an error of plus or minus 0.1 mg/liter at
1.0 mg F/liter.
                 SUBSTANCE

                 Alkalinity

                 Aluminum

                 Chloride

                 Iron

                 Hexametaphosphate

                 Phosphate

                 Sulfate

                 Chlorine
 luoride Determination in
^ater, E.P.A.  Training
"lanual,  National  Training
and Operational  Technology
Center,  Cincinnati, Ohio
 5268
                                7,000 (+)

                                    4 (-)

                               20,000 (-)

                                  200 (-)

                              >50,000

                              >50,000

                               50,000 (-)

                               >5,000
From this, it can be seen that with the use of the
buffer, very few ions normally used in the treat-
ment of water will interfere.

Variations in ionic strength can interfere but again
the make-up of the buffer will adjust this to a
point where interferences are eliminated.

All samples are actually diluted in half by the
addition of an equal volume of buffer solution.
Since this same dilution ratio of equal volumes is
carried out for all samples and standards the
original concentration is used.  Thus if a 1.0 mg/
liter solution is diluted with buffer and is
actually 0.5 mg/liter it is read as 1.0 mg/liter.

Plot the concentration of fluoride in mg/1 on the
log axis and the millivolt reading on the linear
axis.  Figure one has a piece of 2 cycle semi log
paper with the concentration values in the EMP
marked.

The graph will have a negative slope; i.e., there
will be a lowering of the millivolt reading as the
concentration increases.  The line will be straight
from 0.1 to over 100 ppm.
     rii D oo

-------
EFFLUENT MONITORING  PROCEDURE:  Determination of Fluoride in Potable and
                               Waste Waters Using a Selective  Ion Electrode
 Field  &  Laboratory Analysis
                                                                           Section
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 F.2.8a.
A selective ion meter usually has several  scales on
the meter face.  One is divided into 14 equal
divisions is the pH scale; another is divided  into
unequal divisions and usually repeats itself at
least twice.  How it is numbered varies with each
manufacturer.   This scale is the logarithmic scale
and should be  used to read the concentration of
selective ions in mg/1.

Since the units, i.e. how it is numbered,  vary the
concentration  range of the scale can be arbitrarily
set.  This scale is usually repeated; for  example,
if the first number on the left is marked  as .1 and
at the middle  of the scale is 1 and the last number
on the right is 10 this gives two complete
repetitions.  Concentration wise this could mean
that the concentration at center can be adjusted to
read 1.0 mg/1  and at far right as 10.0 mg/1.  This
would then allow concentrations of from 0.1 to 1.0
to 10.0 to be  read directly from the meter face.
                                                                                  Ell.B-29

-------
                                                                    Section  IX
                             TYPICAL LABORATORY  DATA SHEET
    Sample Collections:
         1.   Name of Plant:
         2.   Sampling Location
         3.   Type of Sample
         4.   Date and Time Collected
         5.   Sample Collector
    Sample Analysis:
         1,   Name of laboratory
         2.   Date and Time Collected
         3.   Sample designation 	
         4.   Methods Used
         5.   Was Sample Distilled
         6.   Concentration of F~ determined
             (in mg/liter)	
    Standard Cone.
       mg j^yiHer	                Millivolt reading
         0.0
         0.2
         0.4
         0.6
         0.8
         1.0
         1.2
         1.4
         1.6
         2.0
Ell.B-30

-------
                           A  PROTOTYPE  FOR  DEVELOPMENT  OF
                           ROUTINE  OPERATIONAL  PROCEDURES

                                     for the

                        PRELIMINARY DISTILLATION  PROCEDURE  FOR
                     FLUORIDE ANALYSIS  OF POTABLE AND WASTEWATERS
                                    as  applied  in

                            WATER TREATMENT  FACILITIES
                         WASTEWATER TREATMENT  FACILITIES
                                    and  in the
                        MONITORING  OF  EFFLUENT WASTEWATERS
                National Training and Operational Technology Center
                       Office  of Water  Program  Operations
                      U.S.  Environmental  Protection  Agency
CH.HAL.f.lab.HMP.l .11.77

                                                                             Ell.C-1

-------
WATER MONITORING PROCEDURE;  Preliminary Distillation Procedure for Fluoride
                             Analysis of Potable and Wastewaters

1. Analysis Objectives:

   The learner will  remove interferences to the fluoride analyses by a pre-
   liminary distillation of the sample.

2. Brief Description of  Analysis:

   The sample is added to a previously adjusted acid water mixture.  The
   fluoride is liberated in the strongly acid mixture as the fluoride ion F~.
   This reacts with  the  excess hydrogen  ion H+ to form hydrofluoric acid HP.
   This material leaches silica from glass and forms hydrofluorsilic acid
   HoS,Fg) which distills over as  an aqueous solution.  All  impurities remain

   behind.

3. Applicability of  this procedure:

   a.  Range:   When high-fluoride samples are distilled (greater than 10 mg F /I),
      all  of  the fluoride may not  be distilled over.   Repeat the distillation
      with 300 ml  distilled water.

   b.  Pretreatment:   None required

   c.  Treatment of Interferences:   None  required
                                                                        Ell.C-3

-------
o
I
                                               FLUORIDE ANALYSIS FLOW SHEET
                                                            Potable  Water
                                           Waste Water
                                       Distill
                                                                                           Distill
  Electrode
  Method
SPADNS
Colorimetric
Method
SPADNS
Colorimetric
Method
Automated
Complexone
Method

-------
WATER MONITORING PROCEDURE:
Preliminary Distillation Procedure for Fluoride
Analysis of Potable and Wastewaters
Equipment and Supply Requirements

A. Capital Equipment:

   Water still or other source of distilled water

B. Reusable Supplies:

   1. One adaptor - offset with outer ground glass joint 24/40 at top and inner
      ground glass joint 24/40 at the bottom, with opening with rubber glove
      connector for thermometer.
   1. One burner, natural - gas type (or for other type gas used in lab)
   3. One boiling flask - 1 liter, Pyrex or Kimax, round bottom, with 24/40
      outer ground glass joint
   4. Two clamps - to hold boiling flask and condenser to ring stand.  One
      should be covered with asbestos or fiberglas to withstand heat.
   5, One condenser - 40 cm long, double jacket, outer ground glass 24/40
      joint at top
   6. One connecting tube - with two inner ground glass 24/40 joints
   7. One cylinder - volumetric, 500 ml
   8. One flask - Erlenmeyer, 300 ml
   9. Twelve glass beads - (not hard glass, i.e., Pyrex or Kimax)
  10. One tube grease - silicone stopcock
  11. One ring stand and rod
  12. One thermometer - 0 to 200°C,  24"  long
  13. Tubing,  Tygon - enough to reach water supply and drain

C. Consumable:

   1. Distilled water - about 6 liters
   2. Silver sulfate - crystals (for removal of chloride concentration greater
      than 2000 rng/1)
   3. Sulfuric acid
   4. Detergent
                                                                        Ell .C-5

-------
                                                                                                          Ell.C-6
 WATER MONITORING PROCEDURE:
    Preliminary Distillation  Procedure  for Fluoride
    Analysis  or Potable  and Wastewaters
 OPERATING PROCEDURES
         STEP SEQUENCE
            ION/OPERA:ING GOALS/SPECIFICATIONS
                                                                                         TRAINING
                                                                                        GUIDE NOTES
PRELIMINARY DISTILLATION
A. Sample Collection
B. Equipment Preparation
C. Reagent Preparation
   1. Distilled Water
   2.  Sulfuric Acid
0.  Still  Preparation
   1.  Connect the sti 11
FOR FLUORIDE ANALYSIS:

 1. Collect a minimum of 300 ml
    in a plastic or hard glass
    container.
 1.  Clean all  glassware in
    detergent.
 2,  Rinse with  distilled water,
 1.  Prepare about six (6)
    liters  of distilled water.
    This  must be free fron
    fluoride.

 1.  Concentrated; no pre-
    paration required.
1.  As shown in Fig.  I.

2.  The ground glass  joints
   should be greased with
   silicons stopcock grease.

3.  The thermometer must be
   capable of reading about
   200°C.
                                la.
                                                           Ib.
                                 la.
                                                           2a.
    Polyethylene bottles are preferred to glass
    bottles and are satisfactory provided they have
    not previously contained high-fluoride concen-
    tration solutions.
    No special  requirements are necessary for
    preservation.
                                                                    on:   Tap waters can contain fluoride.
la.  Use a still  or pass tap water through an ion-
    exchange column containing a strongly acidic
    cation exchange resin mixed with a strongly
    basic anion  exchange resin.
    Leave the joint between the adaptor and connecting
    tube open.
    Very small  amounts  are used.   One way to do this
    is to coat the joint then lightly wipe the grease
    off with a soft tissue.

-------
 WATER MONITORING  PROCEDURE:
   Preliminary  Distillation  Procedure  for  Fluoride
   Analysis  of  Potable  and Wastewaters
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Still  Preparation
   (Continued)
   2. Add the Reagents
   3. Close the Still
   4. Heat the Still
1.  Place 400 ml  of distilled
   water in the boiling flask

2.  Carefully add 200 ml con-
   centrated sulfuric acid.

3.  Add about 10 to 12 glass
   beads.
                           4. Shake the flask by swirling
                              to mix the two liquids.
1. Connect the joint between
   the adaptor and the
   connector tube.

2. Check all joints to assure
   tightness.

1. Turn on water to the
   condenser.
la.  Use a 500 ml  graduated cylinder.
2a.  Caution:   Heat is generated, also use protective
    eye covering.

3a.  These serve a  two fold purpose:   first to act as
    boiling stones to prevent superheating of the
    mixed liquids  and secondly to serve as a source
    of silica for  a chemical  reaction in the flask.

      F" + H •+ HF  + Si •+ H,SiFc
                          L   b
    If the glass beads are not present, the silica
    will be taken  from the boiling flask and will
    considerably shorten its  useful  life.

4a.  Hold the ring  stand top and rotate in circular
    fashion.   The  mixing must be complete or the
    liquid can superheat.  If the flask is observed
    closely while  mixing, gradient lines (wavy lines)
    can be seen.  Mix until no more  are seen.

la.  Consult Figure I.
2a. Fluoride will be lost through any joints that
    are not tight.

la. The water should enter at the bottom of the
    condenser.
Ib. The flow of water should be maintained at such a
    rate that condensation does occur in the upper
    third of the condenser.
                                                                                       VIII.2.D.4
                                                                                       (p.  15)
                                                                                                              Ell.C-7

-------
MATER MONITORING PROCfcOURr:  Preliminary Distillation  Procedure  for  Fluoride
                             Analysis  of Potable  and Wastewaters
                                                                                                         Ell.C-8
 OPFRATING  FRCCFDURES
D.  Still  Preparation
   (Continued)
        SUP  SEIJLNCE
2.  Turn on and ignite the
   burner.
                           3.  Place  a  300 ml Erlenmeyer
                              flask  under the condenser,

                           4,  Begin  to heat slowly.

                           5.  When boiling begins, heat-
                              ing nay be increased.
                          6. Continue heating until the
                             temperature reaches 180°C,
                          7. Remove heat.

                          8. Discard distillate.
     INFRMATION/CPLPJTING COALS/SPECIFICATIONS
2a.  A heating mantle may be used;  however,  it is  in-
    convenient to use.   The temperature must be
    stopped at 180°C and it is  hard to do with  a
    mantle.  If a mantle is used,  it can be turned  off
    at a somewhat lower temperature, allowing the
    residual  heat to carry the  temperature  to 180°C.
    How early should be determined by practice  since
    all mantles very.
                               4a.  If  bumping  occurs, mixing  has  not  been  completed.

                               5a.  Adjust  the  flame  to prevent  it  from  contacting  the
                                   distilling  flask  above  the liquid  level.   Super-
                                   heating of  the  vapor  results in high  sulfate
                                   carryover which causes  a sulfate interference.

                               6a.  Distillation must be  stopped when  the temperature
                                   reaches 180°C.  Higher  temperatures  result in
                                   excessive sulfate carryover.
                               6b.  About 45 minutes.

                               7a.  Turn off burner.

                               8a.  Discard this distillate, since  it  contains traces
                                   of  fluoride from the  acid  and glassware.   This
                                   preliminary procedure also serves  to adjust the
                                   acid-water  ratio for  subsequent distillations.
                               8b.  This preliminary distillation need be repeated
                                   only if new acid is used or obvious contamination
                                   occurs or if the operator  has doubts as to
                                   contamination.
                               8c.  The resulting acid-water ratio  can be used over
                                   and over for samples  until  a brown color pre-
                                   dominates and 180°C is  hard to obtain.
                               8d.  The amount of distillate should be 300 ml.
 TRAINING
GUIDE NOTES

-------
i-.!ATER MONITORING PROCEDURL:   Preliminary Distillation Procedure for Fluoride
                              Analysis  of Potable and Wastewaters
 OPERATING  PRCCEDURES
E.  Procedure
   1,  Sample Addition
   2.  Distill  sample
        STEP  SEQUENCE
1, Allow the still  to cool
   until the temperature
   drops to 120°C or lower,

2. Disconnect the joint be-
   tween the adapter and
   connecting tube.

3, Remove high levels of
   chloride in the sample.
                           4.  Add  300 ml  sample.
                           5.  Mix thoroughly.
6.  Close joint between
   adapter and connecting
   tube.

1.  Place a 300 ml  Erlenmeyer
   flask under condenser.

2.  Turn on water to the
   condenser.

3.  Ignite burner.
     IHFCPMATICu/CPlRAYING  GOALS/SPECIFICAFIONS
la. Caution:  The glass is hot; touch with care.
                                                          2a.  The  connecting  tube may be lifted  clear  of  the
                                                              adapter and  condenser.
3a. Add silver sulfate to the sample at the rate of
    5 mg per milligram of chloride when high-chloride
    samples are distilled.
3b. Most potable waters will not have high-chlorides.
    Sea water and brackish waters may have.  As a gen-
    eralized guide, about 2000 rng/1 chlroide should be
    a starting point for adding the silver sulfate.

4a. Use a 500 ml graduated cylinder.
4b. Pour into adapter.

5a. Caution:  Improper mixing may cause super-heating
    and bumping,
5b. Use support ring and swirl in circular manner.
                                                          2a.  If  it  has  been  turned  off.
 TRAINING
GUIDE NOTES
                                                                                                            Ell.C-9

-------
WATER MONITORING PROCEDURE:
   Preliminary Distillation Procedure for Fluoride
   Analysis of Potable and Wastewaters
                                                                                                        Ell.C-10
 OPERATING  PROCEDURES
        STEP SLQUtfXE
    INFORMATION/OPIATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E. Procedure (Continued)
   3. Remove Distillate
4. Begin to heat slowly.

5. Increase temperature.


6. Heat to 180°C.

7. Remove heat.

1. Retain distillate for
   analysis of the fluoride.
F.  Fluoride Determinatior
1.  Chose an approved method.
                          2.  Determine the  fluoride
                             concentration.
4a. Until boiling begins.

5a. Cautiori:  Do not allow flame to go above liquid
    level.



7a. When temperature reaches 180°C.

la. When high-fluoride samples (> 10 mg F~/Hter) are
    distilled, repeat the distillation using 300 ml
    of distilled water.  If substantial amounts of
    fluoride appear in the second distillate, add the
    amount  to that obtained initially and flush
    (300 ml  of water) the still again.  Quantities of
    less than 0.1 mg/1  F~ may be disregarded.
la.  For wastewater the SPADNS-electrode or automated
    complexone methods are approved after
    distillation.
Ib,  For potable water the SPADNS or electrode methods
    are approved.   Distillation is not required for
    the electrode  method.
Ic.  Consult the appropriate EMP on the procedures.

-------
WAiER MONITOR 1NG PROCEPURL:  Preliminary Distillation Procedure for Fluoride
    :                         Analysis of Potable and Wastewaters
 fPERATINr, PROCEDURES
       STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G. Storage

   1. The Still



   2. The Acid
1.  Store the glass still  in
   such a manner as to pre-
   vent physical  damage.

1.  Store in the sti11.
la.  This acid may be reused until  the buildups of
    impurities discolors or prevents the temperature
    from being obtained.
lb.  The time is dependent upon content of impurities
    in the water sample.
Ic.  A good practice would be to occasionally run a
    known standard to assure complete distillation
    of fluoride.
                                                                                                             Ell.C-11

-------
                             CONNECTING TUBE
                             12mm. I.D.
              THERMOMETER
             RUBBER SLEEVE
                                 S 24/40
                                   JOINTS
                          4J   ADAPTER
     1000 ml. BOILING FLASK
                              f 24/40 JOINT
                    BURNER
                                              CONDENSER
                                              300 ml. ERLENMEYER
               FIGURE 1. DISTILLATION APPARATUS
Ell.C-12

-------
WATER MONITORING PROCEDURE:  Preliminary Distillation Procedure for Fluoride
                             Analysis of Potable and Wastewaters
                              TRAINING GUIDE


SECTION                           TOPIC

   I*                   Introduction

  II                    Educational Concepts - Mathematics

 III                    Educational Concepts - Science

  IV                    Educational Concepts - Communications

   V                    Field and Laboratory Equipment

  VI                    Field and Laboratory Reagents

 VII                    Field and Laboratory Analysis

VIII*                   Safety

  IX                    Records and Reports
*Training Guide materials are presented here under the headings marked*.
These standardized headings are used through this series of procedures.
                                                                         Ell.C-13

-------
WATER MONITORING PROCEDURES:
Preliminary Distillation Procedure for Fluoride
Analysis of Potable and Wastewaters
INTRODUCTION
                                             Section  I
                                 TRAINING GUIDE NOTE
                                         REFERENCES/RESOURCES
                 The Guidelines  Establishing  Test Procedures  for the
                 Analysis  of Pollutants  lists three approved  analyti-
                 cal  methods for fluoride.   It 1s mandatory to  pre-
                 cede each by the preliminary distillation  procedure,
                 unless  comparability data  is available on  a  repre-
                 sentative effluent sample  to show that this  pre-
                 liminary  distillation Is not necessary.

                 The National  Interim Primary Drinking  Water  Regu-
                 lations lists two approved analytical  methods  for
                 fluoride.   It is mandatory to precede  only the
                 coloritnetric  (SPAONS) method by  the preliminary
                 distillation procedure.

                 The Preliminary Distillation step required to  be
                 used is the method utilizing a distillation  from
                 suIfuric  acid.
                                        Federal  Register,  Part II,
                                        Wednesday,  Dec,  1,  1976,
                                        page 52782, Parameter 20.
                                        Federal  Register,  Part  iy,
                                        Wednesday,  Dec.  24,  1975,
                                        page 59573, Parameter 10.
                                        Standard  Methods  for  the
                                        Examination  of Water  and
                                        Wastewater.   13th ed.,
                                        page  171.  14th ed.,
                                        page  389.
   Ell.C-14

-------
WATER MONITORING PROCEDURES:
Preliminary Distillation Procedure for Fluoride
Analysis of Potable and Wastewaters
 SAFETY
                                             Section VIII
                                  TRAINING GUIDE NOTE
                                         REFERENCES/RESOURCES
D.Z.4
                 Care should be taken to assure complete mixing
                 between the acid and water or sample.  Incomplete
                 mixing will result in a violent bumping of the still
                 which can throw the acid mixture out of the distill-
                 ing flask and possible damage to the still, analyst
                 and loss of sample.
                                                                               Ell .C-15

-------
                           A PROTOTYPE FOR DEVELOPMENT OF
                           ROUTINE OPERATIONAL PROCEDURES

                                       for the

                           DETERMINATION OF BARIUM  (Ba++)
                                    as applied in

                      WATER AND WASTEWATER TREATMENT FACILITIES
                                     and in the
                          MONITORING OF EFFLUENT WASTEWATERS
                National Training and Operational Technology Center
                       Office of Water Program Operations
                      U.S.  Environmental  Protection  Agency
CH.ME.ba.lab.WMP.l .11.77

                                                                             E12-1

-------
WATER MONITORING PROCEDURE:   Determination of Barium (BA++)


1. Analysis Objectives:

   The learner will  be instructed on how to determine the barium content of a
   sample.

2. Brief Description of  Analysis:

   The sample is first digested with acid to assure all  metals are in a soluable
   form.  After which a  portion of the sample is aspirated into an atomic ab-
   sorption spectrophotometer utilizing an acetylene-nitrous oxide flame.

3. Applicability of the  Procedure:

   The method works for  both potable and wastewaters.

   a. Range of Concentration - The method is recommended for use in the range
      of 1.0 to 20 mg of barium/liter.  The detection limit  is 0.03 mg/1.

   b. Pretreatment of the sample - Digestion in acid pH  to assure solubil ization.
      See Section B.

   c. Treatment of Interferences in the sample - The use of  the nitrous oxide-
      acetylene flame virtually eliminates chemical interferences; however,
      barium is easily ionized in this flame and potassium must be added to
      standards and samples  alike to control this effect.

   d. Source of Procedure -  Manual of Methods for Chemical Analysis of Water
      and Wastes; 1974 ed.,  p. 97; USEPA Technology Transfer, Cinti., OH 45268.
                                                                         E12-3

-------
    WATER MONITORING PROCEDURE;  Determination of Barium (BA++)
    Operating Procedures:
    A. Equipment Preparation
    B. Reagent Preparation
    C. Standard Preparation
    D. Sample Pretreatment
    E, Instrument Calibration
    F. Sample Analysis
    G. Calculations
El 2-4

-------
WATER MONITORING PROCEDURE:  Determination of Barium (BA++)


Equipment and Supply Requirements

A. Capital Equipment:

   1. Atomic absorption spectrophotometer
   2. Recorder - compatible with the instrument
   3. Nitrous oxide burner head
   4. Barium - hollow cathode lamp
   5. Pressure regulators two stage for
      Acetylene - CGA inlet 510 connector
      Air - CGA inlet 1340 connector
      Nitrous oxide - CGA inlet 1320 connector
   6. Balance - analytical with a 0.1  milligram sensitivity
   7. Still - borosilicate glass or equivalent

B. Reusable Supplies:

   1. Trip balance - 100 gram capacity
   2. Pen or pencil
   3. Twelve inch ruler
   4. Hot plate
   5. Five reagent bottles - clear glass, glass stoppered 500 ml  cap
   6. Six beakers - glass, 150 ml size
   7. Six cylinders - graduated, 100 ml size
   8. Three flasks - volumetric, 1000 ml volume
   9. Six flasks - volumetric, 100 ml  volume
  10. One pipet - graduated, 1 ml
  11. Two pi pets - graduated, 10 ml
  12. One pipet - volumetric, 1 ml
  13. Two pipets - volumetric 5 ml
  14. One pipet - volumetric 10 ml
  15. One pipet - volumetric 20 ml
  16. One funnel - powder, glass
  17. One funnel - filtering
  18. Pipet bulb
  19. Safety glasses
  20. Wash bottle - plastic

Consumable Supplies:

   1. Deionizing column - mixed bed type
   2. Gases
      Acetylene - purified or commercial grade
      Air - dry grade
      Nitrous oxide - technical grade
   3. Detergent
   4. Nitric Acid - ACS grade
   5. Hydrochloric acid - ACS grade
   6. Barium chloride
   7. Potassium chloride
                                                                         E12-5

-------
   WATER MONITORING PROCEDURE:  Determination of Barium
    Equipment and Supply Requirements  (Continued)

      8. pH paper - capable of measuring pH 2  (pHydrion)
      9. Filter paper - Whatman #42
     10. Graph paper
     11. Wax marking pencil
     12. Plastic weighing boats, ^ 12 each
E12-6

-------
                                                         + +^
 WATER  MONITORING  PROCEDURE:   Determination of Barium (Ba  )
 OPERATING PROCEDURES
        STEP  StQUENCE
    INFORMATION/OPE RAT ING  GOALS/SPEC IFICATIONS
 TRAINING
GUIDE NOTES
A. Equipment Preparation

   1.  Cleaning of
      Glassware
   2.  Balance Inspection
   3.  Instrument
      Inspection
1. Wash with detergent.



2. Rinse with tap water.

3. Rinse with 1:1 nitric acid
4.  Rinse with tap water.

5.  Rinse with 1:1 hydro-
   chloric acid.

6.  Rinse with tap water.

7.  Rinse with distilled water

1.  Check all  balances for
   cleanliness and proper
   operation.

1.  Check all  items of the
   instrument to  assure
   proper optimization and
   operation.
la. Cleaning should be carried out in this order.
Ib. All glassware should be kept covered after
    cleaning.
                                                           3a.  Add an equal  volume of acid to an equal  volume of
                                                               distilled water.   (Example:  500 ml  of acid added
                                                               to 500 ml of water.
                                                           3b.  Caution:   Always  add acid to water not the
                                                               reverse.
                                                           5a.  Add equal  volume of HC1  to an equal  volume of
                                                               distilled  water.
la.  Consult the manufacturer's  manual  if the  balance
    does not operate properly.
la.  Consult the manufacturer's  manual.
                                                                                                              E12-7

-------
                                                        +-K
 WATER MONITORING PROCEDURE:  Determination of Barium (Ba   )
                                                                                                         E12-8
 OPERATING PROCEDURES
        STEP SEQUENCE
                                 INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE  NOTES
B.  Reagent Preparation

   1.  Distilled
      Deionized Water
   2.  Nitric Acid
      Concentrated
      (HN03)

   3.  Nitric Acid 1:1
   4.  Hydrochloric  Acid
      Concentrated  (HC1)

   5.  Hydrochloric  Acid
      1:1
   6.  Barium Stock
      Solution
1. Prepare by passing dis-
   tilled water through a
   mixed bed of cation and
   anion exchange resins.

1. Commercially available
   reagent grade.
1. Prepare a 1:1 solution  from
   reagent grade nitric acid
   by adding an equal  amount
   of the acid to an equal
   amount of distilled water.

1. Commercially available
   reagent grade.
1
Prepare a 1:1  solution from
reagent grade hydrochloric
acid by adding an equal
amount of the acid to an
equal amount of distilled
water.

Carefully weigh 1.7787
grams of barium chloride
                             la. Use deionized distilled water for the preparation
                                 of all reagents, calibration standards and as
                                 dilution water.
                             la. If a high reagent blank is obtained, it may be
                                 necessary to distill the acid or purchase a
                                 better purity.

                             la. Caution:  add the acid to water; not the reverse.
la.  Caution:   add the acid to water; not the reverse.
                              (BaCl.
         •2H20)
             (analytical
                              reagent  grade).

                           2.  Transfer Into a  1000 ml
                              volumetric  flask.
                                la.  Use a  plastic  weighing boat and  an analytical
                                    balance.
                                2a.  Use a powder funnel.
                                2b.  Use a plastic wash  bottle to rinse weighing boat
                                    and funnel  into the flask.

-------
 WATER  MONITORING  PROCEDURE:  Determination of Barium (Ba1
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFOKMATION/OPERA! If,G  G00L.S/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   7.  Barium Standard
      Solution
   8. Potassium Chloride
      Solution (KC1)
   9. Fuel and Oxidant
3.  Dissolve in distilled
   water.

4.  Dilute to the mark with
   distilled water.

1.  Transfer about 50 ml  of
   distilled deionized water
   to a 100 ml volumetric
   flask.

2.  Pipet 1 ml  of barium stock
   solution (Reagent 6)  into
   the flask.

3.  Dilute to the mark with
   distilled deionized water.

1.  Weigh 95 grams of potassium
   chloride (KC1) (analytical
   reagent grade).

2.  Transfer into a  1000 ml
   volumetric flask.
                           3. Dissolve in distilled
                              water.
4. Dilute to the mark with
   distilled water.

1. Commercial grade  acetylene
   is generally acceptable.
                                                           4a, One ml equals 1  mg Ba (1000 mg/liter)
la.  This volume need not be measured exactly.
                                                           2a. Use a 1  ml volumetric pipet.
3a. This solution contains 10 mg Ba/liter,
    (1  ml = 0.01  mn Ba)

la. Use a plastic weighing boat on a trip balance.
                                                           2a. Use a powder funnel.
                                                           2b. Use a plastic wash bottle to rinse the weighing
                                                               boat and funnel  into flask.
                                                           4a. One ml = 50 mg K = (50,000 mg/liter)
                                                                                                              El 2-9

-------
 HATER MONITORING PROCEDURE:   Determination of Barium (Ba  )
                                                                                                         E12-10
 OPERATING PROCEDURES
        STfP SEQUENCE
                       INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                                      TRAINING
                                                                     GUIDE  NOTES
B. Reagent Preparation
   (Continued)
2. Reagent grade nitrous
   oxide (N~Q) is required.

3. Air - dry grade.
                                                           3a.  Air may be supplied from a compressed air line,
                                                               a compressor or cylinder.
                                                           3b.  Air is used to ignite the burner initially on
                                                               most instruments,  then the flame is  switched
                                                               over to the nitrous oxide.  This reduces the
                                                               possibility of flash back occurring.
                                                           3c.  Caution:   The air  supply must be free from oil
                                                               or other  contaminants.
C. Standard Preparation
1.  Collect 5 clean 100 ml
   volumetric flasks.
                           2.
   Add about
   distilled
   flask.
50 ml
water
deionized
to each
                           3.  Add 0.2 ml  concentrated
                              nitric  acid.
                           4.  Add  5.0  ml  of  concentrated
                              hydrochloric acid.

                           5.  Add  2.0  ml  of  the  potassiun
                              chloride solution  (Reagent
                              8)
2a.  This volume is not critical—just fill  flask
    about half full.
                                3a.  The calibration standards  should  be prepared
                                    using  the same type of acids  (HC1  and  HNO-)  and

                                    at the same concentrations as  the samples  for
                                    analysis.
                                3b,  Nitric acid will  have been added  to the  sample
                                    for preservation.
                                3c.  Use a  1 ml  pi pet  graduated in  tenths.
                                3d.  Use caution when  pipetting concentrated  acids.
                                3e.  Use a  pipet bulb  and safety glasses.

                                4a.  Use a  10  ml  graduated pipet.
                                5a,  Use  a  10 ml  graduated  pipet.
                                                                          VI.C.5
                                                                          (P. 17)

-------
WATER MONITORING  PROCEDURE:   Determination of Barium (Ba
 OPERATING PRCCEDURES
        5T£-  SEQUENCE
    I:NFGKMAT ION/OFF BATING  GOALS/SPEC IFICATIONS
 TRAINING
GUIDE NOTES
C.  Standard Preparation
   (Continued)
D. Sample Pretreatment:
   for Total Metal
6.  Transfer 0.0,  5.0, 10.0,
   15.0, 20.0 ml  of the
   barium standard solution
   (Reagent 6) to each of the
   five volumetric flasks
   respectively.

7.  Dilute each flask to the
   volume mark with de-
   ionized distilled water.

8.  Mix thoroughly.
1. Transfer 100 ml  of
   sample into a clean 150 ml
   beaker.

2. Check the pH.
                           3.  Adjust pH to below 2 if it
                              has not been adjusted
                              during sampling.

                           4.  Add 5 ml  of 1:1  hydro-
                              chloric acid.

                           5.  Heat the sample  at 95°C
                              for 15 minutes.

                           6.  Cool to room temperature.

                           7.  Wash down walls  of the
                              beaker with distilled
                              water.
6a. Use volumetric pipets for the transfer.
6b. The blank (the 0.0 flask) will  not need  to be
    pipetted.
                                                           7a.  The solutions  contain 0.0,  0.5,  1.0,  1.5,  and
                                                               2.0 mg  Ba/liter of solution.
la. Use a 100 ml  graduated cylinder.
2a. pH paper can be used (pHydrion).
2b. pH must be 
-------
 WATER MONITORING PROCEDURE:   Determination of Barium (Ba  )
                                                                                                         EU'-12
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D. Sample Pretreatment:
   for Total Metal
   (Continued)
8. Filter the sample through
   a filter paper into a
   100 ml graduated cylinder.

9. Adjust the volume to
   100 ml.
8a. A paper such as Whatman #42 should be used.
                                                           9a.  The sample is  now ready  for  analysis.
E.  Instrument Calibration
1.  Install  the nitrous oxide
   burner head.
                           2.  Assure that the barium
                              lamp is in  the  light  path.
                           3.  Adjust  the  wavelength  to
                              the  proper  setting.

                           4.  Select  the  proper  slit
                              arrangement,

                           5.  Set  the lamp current con-
                              trol  to its lowest
                              position.

                           6.  Turn  on the power  to the
                              instrument.

                           7.  Adjust  the  lamp current
                              control  until the  proper
                              mill lamps are applied  to
                              the  lamp.
la. Follow the manufacturer's directions for in-
    stallation.  Usually no tools are required and
    the installation is not difficult.

2a. Depending on the instrument, the analyst may
    have to place the lamp into the lamp compartment
    or revolve the turret.

3a. 553.6 nm for barium.
                                4a.  Consult  the manufacturer's  specifications  or
                                    procedures manual.

                                5a.  Consult  the manual  for  location on your
                                    instrument.
                               6a. Consult  the manufacturer's manual  for  location.
                               6b. Allow warm up  time  for  single beam instruments.

                               7a. Consult  the lamp  for proper operating  current.

-------
WATER MONITORING PROCEC'JRF.:   Determination of Barium (Ba  )
 OPERATING PROCEDURES
E.  Instrument Calibration
   (Continued)
F. Sample Analysis
        STLP SEQUENCE
8. Turn on the master valves
   for the air, acetylene
   and nitrous oxide.

9. Ignite flame.
                          10.  Aspirate the standard
                              solution prepared in
                              Section C.

                          11.  Prepare a calibration
                              curve by plotting the
                              concentration, in mg Ba/
                              liter, of the standards
                              against the response for
                              each concentration.
1. Each sample must be pre-
   treated to assure that all
   materials which might
   contain barium are in
   soluble form.

2. Add 2 ml of the potassium
   chloride solution (Re-
   agent 8).

3. Aspirate the unknown
   solution(s) into the
   instrument immediately
   following the  aspiration
   of the standards.
     IfiFCRMATICN/OPEPATING GOALS/SPEC I PI CAT IONS
 8a.  The pressure setting for the gases should be set
     to the manufacturer's recommendations.
 9a.  Using air-acetylene:   after ignition, switch to
     nitrous oxide.
 9b.  Adjust the acetylene  for maximum absorption.
 9c.  See instruction manual  for your particular
     instrument.

lOa.  Record the response on  a recorder or its
     equivalent.
 la.  Steps  are in Section C.
                                                           2a.  Use a 10 ml  graduated pipet.
                                                           3a.  The standard  curve produced under Section E  must
                                                               be verified  each time barium is  to be analyzed.
                                                           3b.  The flame characteristics and instrument
                                                               settings  should  be the same for  standards and
                                                               unknowns.
 TRAINING
GUIDE NOTES
                                                                                                             E12-13

-------
 WATER MONITORING PROCEDURE:   Determination of Barium (Ba  )
                                                                                                         E12-14
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F. Sample Analysis
   (Continued)
G.  Calculations
4.  Record the response.
   Determine the concentra-
   tion of barium in the
   sample by substituting the
   observed instrumental  re-
   sponse on the calibration
   graph.
                           2.
   Express  all  values
   mg  Ba/liter.
                                                 as
la.  Consult the outline in Section F concerning
    "Calibration Graphs."
Ib.  Some instruments can be calibrated to read
    directly in concentration.   This should be used
    only after the analyst is assured that correct
    responses can be attained.

-------
WATER MONITORING PROCEDURE:  Determination of Barium (Ba++)
                               TRAINING GUIDE

SECTION                            TOPIC

   I                      Introduction

  II                      Educational Concepts - Mathematics

 III                      Educational Concepts - Science

  IV                      Educational Concepts - Communications

   V*                     Field and Laboratory Equipment

  VI*                     Field and Laboratory Reagents

 VII*                     Field and Laboratory Analysis

VIII                      Safety

  IX                      Records and Reports
 *Training Guide materials are presented here under the headings marked*.
 These standardized headings are used through this series of procedures
                                                                         £12-15

-------
                                                         •H-,
WATER MONITORING PROCEDURES:   Determination  of  Barium  (Ba   )
 FIELD AND LABORATORY EQUIPMENT
                                                           Section V
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
D.4.1a
B.9.2
As acetylene  (CgH2)  is  packed  dissolved  in  acetone
(CH3COCH3), cylinders should be stored only in an

upright position.  Acetone can be introduced into
the flame if the cylinder has been stored lying on
its side or if used below 75 psig.  Acetone entrain-
ment usually produces a slight pink tinge and ab-
normally high background signals.

The use of a nitrous oxide-acetylene flame is
recommended for the determination of barium.  This
flame virtually eliminates chemical interferences.
However, if nitrous oxide is not available an air-
acetylene flame can be used.
  E12-16

-------
WATER MONITORING PROCEDURES:   Determination of Barium (Ba
FIELD AND LABORATORY REAGENTS
                                                           Section  VI
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
B.8
B.7
C.5
Barium is easily ionized in the nitrous oxide flame.
lonization will produce incorrect results.  In order
to suppress the ionization effect potassium is added
to standard and sample alike.  The concentration in
the standards and samples should equal 1000 mg
K/1000 ml of solution.  For the concentration used
in the preparation, 2 ml of the potassium solution
should be added per 100 ml of standard or sample.

If the nitrous oxide flame is not availabe and the
acetylene-air flame is used, phosphate, silicon
and aluminum will  severely depress the barium
absorbance.   This  may be overcome by the addition
of 2000 mg La/liter.  This solution is prepared as
follows:   Dissolve 58.65 g lanthanum oxide (La?0_)
in 250 ml of concentrated hydrochloric acid
(Caution:  reaction is violent; use a hood).   Dilute
to 500 ml volume.   This solution contains 100 mg La/
ml.  Add  2 ml  of this solution to each 100 ml  of
standard  and/or sample.  This gives a final con-
centration of 2000 mg La/1000 ml of solution.
                                                                               £12-17

-------
WATER MONITORING PROCEDURES:  Determination of Barium  (Ba  )
 FIELD AND LABORATORY ANALYSIS
                                                          Section  VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
 F.3a
In order to comply with the quality control  section
of the "Criteria and Procedures for Water Supply
Laboratory Certification," the minimum requirement
is to run a validation of the standard curve con-
sisting of at least a reagent blank and one  standard
at or near the maximum contaminant level  of
1.0 mg Ba/liter.  This should be done daily  or with
each batch of samples.

If 20 or more samples are run daily, the standard
curve must be verified by running an additional
standard of midrange every 20 samples.

All checks must be within +_ 10 percent of the
original curve.  If not, a new standard curve must
be prepared.

The control recommended for water pollution  analysis
as listed in the Handbook for Analytical  Quality
Control in Water and Wastewater Laboratories is
to verify the standard curve by two standards, one
high and one low concentration.
Criteria and Procedures for
Water Supply Laboratory
Certification.  Office of
Research and Development,
U.S. EPA, Washington, DC
20460.
                                                                     Handbook for Analytical
                                                                     Control in Water and
                                                                     Wastewater Laboratories.
                                                                     U.S. EPA Technology
                                                                     Transfer, Cinti., OH 45268.
   E12-18

-------
                         RESIDUAL CHLORINE  AND  TURBIDITY

I.   INTRODUCTION

    The Interim Primary Drinking  Water  Regulations  (Federal  Register,  December  24,
    1975)  permits  the options  of  substitution of  up to  75  percent  of the  bac-
    teriological  samples with  residual  chlorine determinations.  Any community
    or non-community water  system may avail  themselves  of  this  option  with
    approval  from  the State based upon  results  of sanitary surveys.  Residual
    chlorine  determinations must  be  carried out at  the  frequency of at least
    four for  each  substituted  microbiological sample.

    Since  many  potable water plants  carry out their own microbiological deter-
    minations,  it  will  be necessary  that these  laboratories  be  certified  for
    the bacteriological  parameters.  Residual chlorine  determinations  may be
    carried out by any person  acceptable to the State and  the analytical
    method and  techniques used must  be  evaluated  in some manner to assure that
    reliable  information is obtained.

    Since  the presence of high turbidity can interfere  with  the disinfection
    capability  of  chlorine,  a  maximum allowable limit has  been  set for turbidity
    as follows:

    A.   One turbidity unit  (TU) as determined by  a  monthly average except
        that  five  or fewer  turbidity units  may  be allowed  if the supplier
        of water can demonstrate  to  the State that  the  higher turbidity
        does  not

        1.  Interfere with  disinfection,

        2.  Prevent  maintenance of residual  of  disinfectant  throughout
           distribution system,  or,

        3.  Interfere with  microbiological  determinations.

    B.   Five  turbidity units based on an average  of two consecutive days.

        The Criteria and Procedures  Document for  Water  Supply Laboratory  Certifi-
        cation  suggests  that some quality control guidelines be instituted for
        the residual  chlorine  and turbidity  measurements at  the State  level for
        the purpose  of  ensuring data validity for these critical measurements.

        In response  to  public  comments  regarding  the proposed Primary  Regulations
        (Federal Register,  December  24, 1975) it  is  stated that operators per-
        forming  residual  chlorine and turbidity analyses "....be certified,
        approved,  or at  least  minimally trained to  perform the  analytical tasks
        before  a State  could accept  their analytical determinations...."
  CH.TURB.3.9.77
                                                                         E13.A-1

-------
 II.   RESIDUAL CHLORINE

      Since residual  chlorine analysis  would  be  carried  out  in  "field"  conditions
      or in the small  laboratories  of treatment  plants,  perhaps  by  unskilled
      operators, it is necessary to keep  the  analytical  method  as simple  as
      possible.   For a number of years, operators  had  utilized  the  orthotolidine
      technique in  a kit form to determine  the chlorine  residual.   Recent
      studies  and regulatory guidelines have  dictated  against this  test procedure.
      The acceptable test procedure is-now  the DPD Test  (13th Ed.,  Standard Methods
      for the  Examination of Water  and  Nastewater.  pgs.  129-132), for which kits
      are available" fromat  least two companies  and which  meet  requirements for
      accuracy and  reliability.   These  kits are  capable  of measuring both free
      and combined  chlorine  of which only the free chlorine  is  measured to meet
      compliance requirements.  Kit procedures call  for  a  premeasured single
      powder or tablet reagent added to the test cell  with the  sample and a
      resultant color development measures  by comparison the standardized colors
      within one minute.  Standard^  Methods  includes cautions regarding  temperature
      and pH control  regardfng "th~Ts~test  parameter and this  test procedure, the
      DPD Test,  is  least effected by temperature and the pH  is  adjusted by the
      added reagents.   The only  interfering substance, oxidized manganese, can
      be determined in a preliminary step and compensated  for in the final  test
      value.

III.   TURBIDITY

      Turbidity has long been used  in the water  supply industry for indicating  proper
      operational techniques. Turbidity  should  be clearly understood to  be an  ex-
      pression of the optical property  of a sample which causes  light to  be scattered
      and absorbed  rather than transmitted  in straight lines through the  sample.

      The standard  method for the determination  of turbidity has been based on  the
      Jackson  candle turbidimeter.   However,  the lowest  turbidity value which can
      be measured directly on the Jackson turbidimeter is  25 units  which  is well
      above the monitoring level.  Because  of these low  level requirements, the
      nephelometric method was chosen and procedures are given  in StandardMethods
      (13th Ed., 1971).

 IV.   NEPHELOMETRIC MEASUREMENTS FOR COMPLIANCE  MONITORING

      The subjectivity and apparatus deficiencies  involved in visual methods of
      measuring turbidity make each unsuitable as  a standard method.

      Since turbidity is an  expression  of the optical  property  of scattering or
      absorbing light, it was natural that  optical  instruments  with photometers
      would be developed for this measurement.
                                                               (1 6)
      The type of equipment  specified for compliance monitoring* '' utilizes
      nephelometry.

      A.  Basic Principle^7'

          The  intensity of light scattered  by the  sample is  compared  (under defined
          conditions) with the intensity  of light  scattered  by  a standard reference
          solution  (formazin).  The greater the  intensity  of scattered  light, the
          greater the turbidity. Readings  are made and  reported in NTUs  (Nephelometric
          Turbidity Units).


 E13.A-2

-------
B.  Schematic
                 Lamp

                  I
                 LJ
                              Meter
                          Photoceli(s)	fC
ns

1
1 	 r~
-1


/"''


, i




f •





furbidity P
^v/Scallei tig
\ «^


**
^
                      Sample Cell
                      (Top View)
Figure 2  NEPHELOMETER

           (90° Scatter)
    Light  passes  through  a  polarizing lens  and  on  to  the  sample  in  a  cell.
    Suspended  particles  (turbidity)  in the  sample  scatter the liqht.

    Photocell (s)  detect  light  scattered by  the  particles  at  a 90° angle  to  the
    path of  the  incident  light.   This light energy is converted  to  an electric
    signal for the meter  to measure.

    1.  Direction of  Entry of  Incident Light to Cell

       a.   The  lamp  might be  positioned as shown in the schematic so the
            beam  enters  a sample  horizontally.

       b.   Another instrument design has the light beam  entering the sample
            (in a flat-bottom  cell)  in a vertical  direction  with the  photocell
            positioned accordingly at a 90° angle  to  the  path of incident light,

    2.  Number of Photocells

       The  schematic shows the photocell(s)  at one 90° angle to the  path of
       the  incident  light.  An instrument  might utilize  more than  one photo-
       cell position, with each  final  position being at  a 90° angle  to  the
       sample liquid.

    3.  Meter  Systems

       a.   The meter might measure  the signal  from the scattered light  in-
            tensity only.

       b.   The meter might measure  the signal  from a ratio  of the  scattered
            light versus light transmitted  directly through  the sample to a
            photocell.
                                                                       E13.A-3

-------
          4.  Meter Scales and Calibration

              a.  The meter may already be calibrated In NTLJs.  In this case,
                  at least one standard is run in each instrument range to be
                  used in order to check the accuracy of the calibration scales,

              b.  If a pre-calibrated"scale is not supplied, a calibration curve
                  is prepared for each range of the instrument by using appropriate
                  dilutions of the standard turbidity suspension.

      C.  EPA Specifications for Instrument Design^ '

          Even when the same suspension is used for calibration of different
          nephelometers, differences in physical design of the turbldimeters will
          cause differences in measured values for the turbidity of the same sample.
          To minimize such differences, the following design variables have been
          specified by the U. S. Environmental Protection Agency.

          1.  Defined Specifications

              a.  Light Source

                  Tungsten lamp operated at not less than 85% of rated voltage
                  and at not more than rated voltage.

              b.  Distance Traveled by Light

                  The total of the distance traversed by the incident light plus
                  scattered light within the sample tube should not exceed 10 cm.

              c.  Angle of Light Acceptance of the Detector

                  Detector centered at 90° to the incident light path and not to
                  exceed ± 30° from 90°.

                  (Ninety degree scatter is specified because the amount of scatter
                  varies with size of particles at different scatter angles).

              d.  Applicable Range

                  The maximum turbidity to be measured is 40 units.  Several ranges
                  will be necessary to obtain adequate coverage.  Use dilution for
                  samples if their turbidity exceeds 40 units.

          2.  Other EPA Design Specifications

              a.  Stray Light

                  Minimal stray light should reach the photocell(s) in the absence
                  of turbidity.
E13.A-4

-------
3.
        Some causes of  stray  light  reaching  the  photocell(s)  are;

        1)  Scratches or  imperfections  in  glass  cell  windows.

        2)  Dirt,  film  or condensation  on  the glass.

        3)  Light  leakages  in  the instrument system.

            A schematic of  these causes  is shown in  Figure 3.
                                            Photocell!*)

                                            Light laokage from
                                            Tronimitted tight
                          Light Scatter by glass lube
                              (Top View)
                     Figure 3   NEPHELOMETER
                     SOURCES OF STRAY LIGHT

        Stray light error  can be as  much  as 0.5  NTU.   Remedies  are
        close inspection of sample cells  for  imperfections  and  dirt,
        and good design which can minimize the effect of stray  light
        by controlling the angle at  which it  reaches  the sample.

b.  Drift

    The turbidimeter should be free  from  significant  drift  after  a
    short warm-up period.  This  is imperative if the  analyst is
    relying on a manufacturer's  solid  scattering standard for setting
    overall instrument sensitivity for all ranges.

c.  Sensitivity

    In waters having turbidities  less  than one unit,  the instrument
    should detect turbidity differences of 0.02  unit  or less.
    Several ranges will be necessary to obtain sufficient sensitivity
    for low turbidities.

Examples of instruments meeting  the  specifications  listed in 1  and 2
above include:

a.  Hach Turbidimeter Model 2100 and  2100A.

b.  Hydroflow Instruments  DRT 100, 200, and 1000.
                                                                 E13.A-5

-------
          4.   Other  turbidimeters meeting  the  listed  specifications are also
              acceptable.
      D.   Sources  of Error
          1.   Sample Cells
              a.   Discard scratched or  etched  cells.
              b.   Do not touch cells where light strikes them in instrument,
                                                               fo\
              c.   Keep cells  scrupulously  clean, inside and out.  '
                  1)  Use detergent solution.
                  2)  Organic solvents  may also be used.
                  3)  Use deionized water  rinses.
                  4)  Rinse and dry with alcohol or acetone.
          2.   Standardizing Suspensions^ '
              a.   Use turbidity - free  water for preparations.  Filter distilled
                  water through a 0.45um pore  size membrane filter if such filtered
                  water shows a lower turbidity than  the distilled water.
              b.   Prepare a new stock suspension of Formazin each month,
              c.   Prepare a new standard suspension and dilutions of Formazin
                  each week,
          3.   Sample Interferences
              a.   Positive
                  1)  Finely  divided air bubbles
              b.   Negative
                  1)  Floating debris
                  2)  Coarse  sediments  (settle)
                  3)  Colored dissolved substances
                     (absorb light)
E13.A-6

-------
    E.  Reporting Results^ '
                NTU                        RECORD TO NEAREST
                0.0-1.0                           0.05
                  1-10                            0.1
                 10-40                            1
                 40-100                           5
                100-400                          10
                400-1000                         50
                   >1000                        100
    F,  Precision and Accuracy'  '
        1.   In a single laboratory fEMSL), using surface water samples at
            levels of 26, 41, 75 and 180 NTU, the standard diviations were
            +0.60, +0.94, +1.2 and +4.7 units, respectively.
        2.   Accuracy data is  not available at this time.
V.   STANDARD SUSPENSIONS AND RELATED
    One of the critical  problems in measuring turbidity has been to find a
    material  which can be made into a reproducible suspension with uniform sized
    particles.  Various  materials have been used.
    A.   Natural  Materials
        1.   Diatomaceous earth
        2.   Fuller's earth
        3.   Kaolin
        4.   Naturally turbid waters,
        Such  suspensions are not suitable as reproducible standards because
        there is no way  to control  the size of the suspended particles.
    B.   Other materials
        1 .   Ground glass
        2.   Microorganisms
        3.   Barium Sulfate
        4.   Lates spheres
        Suspensions of these also proved inadequate.
                                                                           E13.A-7

-------
         Formazin

         1.  A polymer formed by reacting hydrazine sulfate and hexamethylenete-
             tramine sulfate.

         2.  It is more reproducible than previously used standards.   Accuracy
             of + one percent for replicate solutions has been reported.

         3.  In 1958, the Association of'Analytical Chemists initiated a  standard-
             ized system of turbidity measurements for the brewing industry by:

             a.  Defining a standard formula for making stock Formazin solutions
                 and

             b   Designating a unit of measurement based on Formazin,  i.e.i the
                 Formazin Turbidity Unit (FTU).

         4.  During the 1960's Formazin was increasingly used for water quality
             turbidity testing.   It is the currently recognized standard  for
             compliance turbidity measurements.
     D.   Units
         1.   At first results were translated into Jackson Turbidity Units (JTU).
             However, the JTU was derived from a visual  measurement using con-
             centrations (mg/liter)  of silica suspensions prepared by Jackson.
             They have no direct relationship to the intensity of light scattered
             at 90 degrees in a nephelometer.

         2.   For a few years, results of nephelometric measurements using specified
             Formazin standards were reported directly as Turbidity Units (TUs).

         3.   Currently, the unit used is named according to the instrument used for
             measuring turbidity.  Specified Formazin standards are used to calibrate
             the instrument and results are reported as Nephelometric Turbidity
             Units (NTUs).
VI.  SUMMARY
     The importance of residual  chlorine determination can be seen in its possible
     effect on the health of the consumers.   The Criteria and Procedures for
     Laboratory Certification suggests that  some form of quality assurance should
     be instituted on a state level  to assure valid data for both the chlorine and
     turbidity measurements.  The comments on the public responses to the proposed
     Interim Primary Regulations also suggests some form of quality assurance on
     the state level to be instituted.  Consequently, the Regional Certification team
     should point out to the principal laboratories the importance of some kind of
     effort being instituted.  States might  wish to offer some kind of formal
     training effort as part of the  approval mechanism for the operators doing
     the chlorine and/or turbidity measurements.
 E13.A-8

-------
                            A PROTOTYPE FOR DEVELOPMENT OF
                            ROUTINE OPERATIONAL PROCEDURES

                                       for the

                 MEASUREMENT OF FREE CHLORINE UTILIZING THE DPD KIT
                                     as applied in

                          DRINKING WATER TREATMENT FACILITIES
                                       and in the
              DISTRIBUTION SYSTEMS OF DRINKING WATER TREATMENT FACILITIES
                 National Training and Operational Technology Center
                          Office of Water Program Operations
                         U.S. Environmental Protection Agency
CH.HAL.cl.lab.WMP.1.11.77
                                                                            E13.3-1

-------
WATER MONITORING PROCEDURE:   Measurement of Free Chlorine Utilizing
                             the DPD Kit


1, Analysis Objectives:

   The actual  use of a kit form of the DPD method for chlorine is  so simple
   a procedure, it need  not  be written.   However, the permanent standards,
   whether liquid or solid,  must be calibrated.   This unit sets down a  method
   that can be used to accomplish this.   The method is applicable  to drinking
   waters only for compliance purposes.

2. Brief Description of  Analysis:

   The kit usually contains  a comparator which is a holder of sorts  to  support
   the sample cell and the standard in such manner as to allow the operator  to
   see both colors and compare their intensities.  The operator adds a  single
   reagent to the water  to be analyzed.   The color is formed  immediately  and
   a comparison is made  between the sample and standards.

3. Applicability of this Procedure:

   a.  Range of concentration - 0.0 to  about 3.0  mg chlorine/liter  or whatever is
      the highest standard supplied with the kit.
   b.  Pretreatment - the sample may not  be preserved;  it should be run  as soon
      as possible or within  one hour after being taken.
   c.  Treatment of Interferences  in Samples:   The procedure includes instructions
      for the  determination  of the interference  caused if oxidized manganes  is
      present.   Other interferences are  suppressed by the make-up  of the  reagents,
Source of this procedure:   Standard Methods  for the Examination  of  Water
and Wastewater, 13th ed.,  pp.  129-132 or 14th ed.  pp.  329-334.
                                                                       E13.B-3

-------
     WATER MONITORING PROCEDURE:   Measurement of Free Chlorine Utilizing
                                  the DPD Kit
     Equipment and Supply Requirements

     A.  Capital Equipment:

        Analytical balance capable of weighing to 0.1  mg (.0001  g)  under
        a 200 g load.

     B.  Reusable Supplies:

        1. Chlorine Test Kit {DPD method)  such as:
           a. Model CN-66
              Hach Chemical Company
              PO Box 907
              Ames, IA 50010

           b. Model LP-1
              LaMotte Chemical Products Company
              Cbestertown, MD 21620

        2. Eight beakers, 50 ml size
        3. Two flasks, volumetric with stoppers, 1000 ml size
        4. Ten flasks, volumetric with stoppers, 100 ml size
        5. One pipet, volumetric, 100 ml size
        6. One pipet, volumetric, 50 ml size
        7. One pipet, volumetric, 20 ml size
        8. Three pipets, volumetric, 10 ml size
        9. One dropper bottle, 100 ml size
       10. Two rubber stoppers - to fit comparator cells {if stoppers are not
           supplied by manufacturer)

     C,  Consumable Supplies:

        1. Distilled water, about 1 gal, or 3.8 liters
        2. Graph paper, arithmetic, 10 x 10 divisions
        3. Pencil or pen
        4. Wash bottle, plastic squeeze type
        5. Weighing boat, plastic, disposable
        6. Potassium Permanganate (KMnO.)
        7. Potassium Iodide Crystals  (KI)
        8, Sodium Arsenite  (NaAsQ-)

        9. DPD - Reagent (Usually a supply is provided with the kit.)
E13.B-4

-------
U'ATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing the DPD Kit
 OPERATING PROCEDURES
        STEP SFQUENCt
    IilF(;RMATION/OPESATIf,5 COALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Reagent Preparation

   1.  Stock Potassium
      Permanganate
      Solution
1.  Weigh out 0.891  grams of
   potassium permanganate
   (KMN04).

2.  Transfer  the potassium
   permanganate to a 1000 ml
   volumetric flask,

3.  Wash the  weighing boat
                           4. Pour the washing into the
                              volumetric flask.

                           5. Repeat washing two more
                              times.
                           6. Add enough water to fill
                              the flask about half full.

                           7. Swirl  until the potassium
                              permanganate is dissolved.

                           8. Dilute to the mark on the
                              volumetric flask.
                           9,  Stopper and mix by re-
                              peated inversions.
la.  Use an analytical  balance.
lb.  Use a plastic weighing boat.
                                                           3a. With distilled water.
                                                           3b. Use a plastic squeeze type wash bottle.
                                                           3c. The wash should be about 10 to 20 ml.
                                5a.  The idea is to assure complete transfer of the
                                    potassium permanganate into the volumetric flask,
                                    Since potassium permanganate has a strong color,
                                    wash until  no color is seen in the weighing boat
                                    containing  water.
                                7a.  If a  magnetic stirrer is available,  this can be
                                    used  to save time.

                                8a.  There is a single ring etched around the neck of
                                    the flask.  When the flask is filled to this
                                    mark, it will contain the volume stated on the
                                    flask.
                                                                                                            L13.B-5

-------
 WATER  MONITORING PROCEDURE:   Measurement of Free Chlorine Utilizing the DPD Kit
                                                                                                        E13.B-6
 OPERATING PROCEDURES
        STEP SEQUENT
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Reagent Preparation
   (Continued)

   2.  Intermediate
      Potassium
      Permanganate
      Solution
   3.  Standard  Potassium
      Permanganate
      Solution
   4. Potassium Iodide
     Crystals

   5. Sodium Arsenite
     (NaAs02)
 1. Transfer  10.00 ml of the
   stock potassium permanga-
   nate solution into a 100ml
   flask.

 2. Dilute to the mark with
   distilled water.

 3. Stopper and mix by re-
   peated inversions.

 1. Transfer  10.00 ml of the
   intermediate potassium
   permanganate solution to a
   1000 ml volumetric flask.

 2. Add distilled water to the
   flask to the mark.

 3. Stopper and mix by re-
   peated inversions.

 1. No preparation  necessary.
1.  Weigh out 0.500 grams of
   sodium arsenite.
                          2. Transfer the sodium
                             arsenite to a 100 ml
                             volumetric flask.
 la. Use a  10.0 ml volumetric pipet.
                                                           2a. Add water  from a  plastic  squeeze bottle.
                                                          3a. This solution has a chlorine equivalent of
                                                              1.00 mg/1 in the DPD reaction.
la.  Use a trip balance or an analytical  balance.
Ib.  Use a plastic weighing boat.
Ic.  CAUTION:   TOXIC - take care to avoid injestion.

-------
 MATER MONITOR!N6 PROCEDURE:   Measurement of Free Chlorine Utilizing the DPD Kit
 OPERATING PROCEDURES
        STEP SEQUENCE
    I'lFCRMATION/OPEaATING COALS/SPECIFICATIONS
 TRAINING
GUIDL NOTES
A. Reagent Preparation
   (Continued)
    6. Preparation of the
       Standard Series
3. Rinse the weighing boat.
                           4, Add the rinses to the
                              volumetric flask.

                           5. Add enough distilled water
                              to fill the flask about
                              half full.

                           6. Swirl until the sodium
                              arsenite is dissolved.

                           7. Dilute to the mark on the
                              volumetric flask.
8. Stopper and mix by re-
   peated inversion.

9. Transfer to a dropper
   bottle, and label.

1. Prepare a series of dilu-
   tions of the standard
   potassium permanganate
   solution in 100 ml
   volumetric flasks.
3a. With distilled water.
3b. Use a plastic squeeze bottle.
                                7a. There is a single ring etched around the neck of
                                    the flask.  When the flask is filled to this
                                    mark, it will contain the volume stated on the
                                    flask.
                                                           9a. Label "Sodium Arsenite."
la.  The series should contain several concentrations
    below and above the expected values.
Ib.  The expected value for drinking water would be
    near 0.2 mg/1  of free chlorine.
Ic.  The series can be prepared as follows:
                                                                                                              E13.B-7

-------
                                                                                                        E13.B-8
WATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing the DPD Kit
OPERATING PROCEDURES
A. Reagent Preparation
(Continued)












B. Calibration of the
Kit Standards











STEP SEQUENCE














1. Fill the reference tube
with distilled water.

2. Rinse the sample cell with
the water to be tested.
3. Fill the cell with the
water to be tested to the
mark.
4. Add the manufacturer's
DPD reagent.



INFORMATION/OPERATING GOALS/ SPECIFICATIONS
Col. 1

ml of
Std. KMnO.

0
10
20
30
40
50
70
100
Col. 2

ml of
Water

100
90
80
70
60
50
30
0
Col. 3
Cone.
Cl"
mg/I

0.0
0.1
0.2
0.3
0.4
0.5
0.7
1.0
Col
. 4
Comparator
Values
mg C
a









r/i
b








Col. 5
*~ Avg. Value
a + b
2 mg/1










la. Not all kits have a reference tube.
Ib. Follow the manufacturer's directions for per-
forming a test with their comparator.
2a. In this case, one of the standards.
2b. Read from low to high on the
3a. Most kits have a line around
series.
the tube. This
indicates a calibrated volume. The cells should
always be filled to this mark.
4a. This is usually in the form of a powder or a
tablet.

4b. This reagent contains both the DPD indicator
and the buffer as prescribed
in the DPD method
In the "Standard Methods" references.
TRAINING
GUIDE NOTES




























-------
WATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing the DPD Kit
 OPERATING  PROCEDURES
        STEP  SEQUEhCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Calibration of the
   Kit Standards
5.  Stopper the cells.



6.  Mix by inversion.

7.  Place cell  in comparator.

8.  Hold comparator up  to the
   light.

9.  Obtain a match between
   color in cell and
   standard.
                          10.  Record the value obtained
                              from the comparator.

                          11.  Discard the solution.
                          12.  Rinse the cell  with the
                              next standard to be run.

                          13.  Discard.

                          14.  Fill to the line with the
                              next standard.

                          15.  Repeat steps 4  through 11
                              for al1 standards.

                          16.  Rinse the tube  with  the
                              first standard  read.
 5a.  Most manufacturers provide stoppers.   If not,
     procure two rubber stoppers.   The diameter of
     the tubuler cells varies so check the size needed

 6a.  One inversion will do.
 8a.  Rotate disc and/or compare color in tube with
     the standard colors.

 9a.  Where the comparator  has only specific values
     such as 0.2, 0.4, 0.6,  etc.  If the color is not
     an exact match the mid-point value between the
     two closest colors is used.

lOa.  Along with the known  concentration
                               16a.  The entire series  should be read again  to obtain
                                    a second value.
                               16b.  The values  should  be  recorded  in Column 4  of
                                    Table  A.6.1c.
                                                                                                             E13.B-9

-------
 WATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing the DPD Kit
                                                                                                         E13.B-10
 OPERATING PROCEDURES
                                STEP  SEQUENCE
                                     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE  NOTES
Calibration of
Kit Standards
(Continued)
                  the
C. Interference
   Determination
17. Fill with solution.

18. Add the reagent.

19. Obtain a value.

20. Repeat steps 16-19 for the
    remaining standards.
17a.  Do not add more water or standard.
                                                          19a. Steps 5 through 9.

                                                          20a. By comparing standards of known concentrations to
                                                               the sealed/permanent visual  standards and plot-
                                                               ting a comparison on graph paper, a correction
                                                               factor can be derived and applied to all  future
                                                               results obtained on the now calibrated apparatus.
                                                          ?0b. This calibration should be carried out at least
                                                               every 6 months and checked by running one or two
                                                               concentrations whenever new reagent (powder or
                                                               pill) is purchased.
                        1.  Fill  the  reference  tube
                           with  the  clear water  to be
                           tested.

                        2.  Rinse the sample  tube with
                           the water to be tested.

                        3.  Discard this rinse.

                        4.  Fill  the  cell to  the mark
                           with  sample.

                        5.  Add one small crystal of
                           potassium iodide.

                        6.  Add one drop of sodium
                           arsenite  solution.

                        7.  Stopper the tube and mix
                           by inverting several
                           times.
                                 la.  This  procedure  must  be carried  out until  it  is
                                     determined  that the  interference  is not present,
                                     then  this section  may  be  omitted.
                                                           4a.  Do  not overfill;  the  reagents  are based  on  this
                                                               volume.
                                                           6a.  Reagent  5.

-------
 HATER MONITORING PROCEDURE:   Measurement of Free Chlorine Utilizing the DPD Kit
 OKRATING PROCEDURES
         STEP
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Interference
   Determination
   (Continued)
D.  Sample Analysis
 8. Unstopper and add the
    manufacturer's reagent.

 9. Stopper and mix by
    inversion.

10. Place cell  in comparator.

11. Hold comparator up to the
    light.

12. Match any red color in the
    sample with the standard.

13. Record the  value.

14. Subtract this value from
    the value obtained in
    section D.
 1.  Fill  the reference tube
    with  the clear water to be
    tested.

 2.  Rinse the sample tube with
    the water to be tested.

 3.  Discard  this rinse.

 4.  Fill  with water to be
    tested.

 5.  Add the  reagent.

 6.  Stopper  and  mix by
    inversion.
8a. This is the DPD reagent and buffer.
                                                          14a.  Step D.10.
la.  If a reference tube is used in the kit.
                                                           4a.  Fill  to the mark on the tube.   Do not over fill;
                                                               the reagents are based on this volume.
                                                                                                             E13.B-11

-------
 MATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing the DPO Kit
                                                                                                         E13.B-12
 OPERATING PROCEDURES
0. Sample Analysis
   (Continued)
E.  Calculations
         STEP SEQUENCE
 7. Insert into the comparator

 8. Hold the comparator up to
    the light.

 9. Match the color developed
    in the sample with the
    standard colors.

10. Check the calibration
    graph.
 1.  Average the values  ob-
    tained  for each standard.

 2.  Record  averages in
    Column  5.

 3.  Plot  the known  value
    (Column 3)  vs.  the  average
    in  Column  5.
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                          IQa.  If no deviation has been noted, the concentration
                                                               can be determined directly from the comparator.
la.  Use the formula at the top of Column 5 in
    Table A.b.lc.
                                                           3a.  Use an  arithmetic paper.
                                                           3b.  See example paper attached.
 TRAINING
GUIDE NOTES

-------
                                                                                                                  c
                                               \- -
                                                                                                     	 _100
i /  i ">  6m IQ  • i
                                            oo
                                                                                                                   "  O
                                                                                                                     ^*
                                                                                                                     ^
                                                                                                                • -CNJ

-------
   WATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing
                                the DPD Kit
                                      DPD KITS
   The DPD N.N-diethyl-p-phenylene diamine method was chosen because it is a
   relatable test and there is a test kit for it.  It is specifically mentioned
   in the comments to the Interim Primary Drinking Water Regulations (p. 59580),
   The kit can be used for many of the forms of chlorine.  However, the only
   form which must be measured and reported is the free chlorine.  The kit
   requires the addition of a solid reagent to the sample, mixing and comparison
   of the reading with standard colors, the color generated by the free chlorine
   and the DPD reagent is formed immediately.

   Any kit using the DPD reagent is acceptable for the measurement.  Both the
   Hach Chemical Company and the LaMotte Chemical Company make kits that can
   be used.

   The permanent color standards provided with the kit should be calibrated
   initially and thereafter periodically to assure correct readings are obtained,

   These kits are not to be used to monitor chlorine under the permit system
   for the National Pollutant Discharge Elimination System (NPDES).  This kit
   is approved only for analysis of chlorine in drinking waters under the
   National Interim Primary Drinking Water Regulations.
E13.B-14

-------
WATER MONITORING PROCEDURE:  Measurement of Free Chlorine Utilizing
                             the DPD Kit
                                 TRAINING GUIDE


SECTION                              TOPIC

   I                    Introduction

  II                    Educational Concepts - Mathematics

 III                    Educational Concepts - Science

  IV                    Educational Concepts - Communications

   V                    Field and Laboratory Equipment

  VI                    Field and Laboratory Reagents

 VII*                   Field and Laboratory Analysis

VIII                    Safety

  IX                    Records and Reports
*Training Guide materials are presented here under the headings marked*.
These standardized headings are used through this series of procedures.
                                                                        F13.B-15

-------
WATER MONITORING PROCEDURES:
              Measurement of Free Chlorine Utilizing
              the DPD Kit
 FIELD AND LABORATORY ANALYSIS
                                                          Section  VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
C.I.la
The only interfering substance likely to be en-
countered 1n water is oxidized manganese.  Conse-
quently an operator should determine If this
interfering substance is present or not.  If not,
the procedure may be omitted.  If it is present,
the procedure must be carried out prior to each
determination.
   E13.B-16

-------
                            A PROTOTYPE FOR DEVELOPMENT OF
                            ROUTINE OPERATIONAL PROCEDURES

                                       for the

                              DETERMINATION OF TURBIDITY
                                     as applied in

                            WATER AND WASTEWATER FACILITIES
                National Training and Operational Technology Center
                       Office of Water Program Operations
                      U.S. Environmental Protection Agency
CH.TURB.lab.WMP.1.11.77
                                                                            E13.C-1

-------
 WATER MONITORING PROCEDURE:  Determination of Turbidity


 1. Analysis Objectives:

   The  user of  the attached material will learn how to use a nephelometric
   type  turbidimeter to measure the turbidity of a sample.

 2. Brief Description of Analysis:

   The method is based upon a comparison of the intensity of light scattered
   by the sample under defined conditions with the intensity of light scattered
   by a  standard reference suspension.  The higher the intensity of scattered
   light, the higher the turbidity.

   A person using this outline should have the basic skills used in a laboratory
   such as:

   a. Preparation of standards
   b. Use of volumetric glassware

 3. Applicability of this Procedure:

   a. Range - From 0 to 40 nephelometric turbidity units or with appropriate
      dilutions higher than 40 units.
   b. Pretreatment of Samples - The sample should be run as soon as possible.
      Preservation is not recommended.
   c. Treatment of Interferences - (1) Floating debris should be removed before
      analysis, (2) air bubbles should be allowed to dissipate before reading,
      (3) with finished waters no other interferences are noted.
Source of Procedure:  Methods for Chemical  Analysis of Water and Wastes, 1974,
Environmental  Protection Agency, Office of Technology Transfer, Cincinnati,
Ohio 45268.  Page 295.

                                                                        E13.C-3

-------
    WATER MONITORING PROCEDURE:   Determination  of Turbidity

    Equipment or Supply Requirements

    A.  Capital Equipment:

       1. Turbidimeter, Nephelometrie - see list of acceptable instruments  meeting
          the following criteria:

          a. Light source - Tungsten  lamp operated at not less than BS% of  rated
             voltage or more than rated voltage.
          b. Distance traveled by incident light and scattered light within the
             sample tube:   total  not  to exceed  10 cm.
          c. Angle of light acceptance of the detector:   centered at 90° to the
             incident light path and  not to exceed + 30° from 90°.
          d. Maximum turbidity to be  measured:   40 units.
       Acceptable Instruments*

        - Hach  Model 2100
                Model 2100 A
        - HF Instruments
Model DRT- 15
         -100
         -150
         -200
Hach Chemical Co., 713 S.  Duff Ave.,
PO Box 907, Ames, IA 50010

HF Instruments, ltd.*  105 Healey Rd.,
Bolton, Ontario, Canada
        - Turner  Model 40-002 (for drinking water)  Turner Designs, 2247A Old Middlefield
                        40-005 (for waste waters)    Way, Mountain View, CA 94043

        • Bausch & Lomb - An attachment for their "Spectronic mini 20" Spectrophotometer.
          This can be obtained from any company that sells Bausch and Lomb
          Spectrophotometers,

        2. Trip balance  (or platform) or analytical balance:  with 0.01 gram
          sensitivity
        3. Distillation  equipment - all glass still or ion exchange cartridges
        4. Standard Turbidity Suspensions  (optional) - if none supplied with the
          Instrument

     B.  Reusable Supplies:

        1. One brush, bottle
        2. One flask, side arm  filtering,  500 ml size
        3. Six flasks, volumetric, 100 ml  size with stoppers
        4. One funnel, membrane filter funnel and holder
        5. Two pi pets, volumetric, 5 ml size
          Three pipets, volumetric, 10 ml size
        6. Pipet bulb
        7. Wash bottle,  squeeze type, 500  ml


     +Also sold by:  Fisher Scientific Co., 711 Forbes Ave., Pittsburgh, PA 15219

     *This list is not meant to be complete.  It covers those known by the author
     at  the time of writing this material.
E13.C-4

-------
WATER MONITORING PROCEDURE:   Determination of Turbidity
C. Consumable Supplies:

   1.  Distilled water
   2.  Detergent
   3.  Membrane filters,  0.45 micron pore size
   4.  Tissues
   5.  Weighing boats, plastic disposable, about 12
   6.  Reagents

      Hexamethylenetetramine, reagent grade - can be purchased from:

      J.  T.  Baker Chemical  Co.      Cat.  No. N145
      Fisher Scientific             Cat.  No. H289
      MC/B or Sargent-Welch        Cat.  No. HX-0280
      A.  H.  Thomas Co.              Cat.  No. C389

      Hydrazine Sulfate, Reagent grade - can be purchased from:

      J.  T.  Baker                  Cat.  No. 2177
      Fisher Scientific             Cat.  No. H-320
      MC/B or Sargent-Welch        Cat.  No. HX-0575
      A.  H.  Thomas Co.              Cat.  No. C393
                                                                        E13.C-5

-------
MA1ER MONITORING PROCEDURE:  Determination of Turbidity
                                                                                                         E13.C-6
OPERATING PROCEDURES
A. Glassware
1. Volumetric





2. Cells









B. Sample Pre-treatment



STEP SEQUENCE

1 . Wash with brush and
detergent.


2. Rinse with tap water.
3. Rinse with turbidity free
water.
1. Rinse with tap water.





2. Rinse with turbidity free
water.
3, Handle and dry with soft
tissue.
1. Measure turbidity as soon
as possible after
sampl ing.

INFORMATION/OPERATING GOALS/SPECIFICATIONS

la. The glassware used for the formaline preparation
may have some polymer adhering to the glass.
Consequently, the brush should be used to clean
this off.



la. Use care in handling the cells at all times Do
not touch them where the light strikes them.
Ib. Consult the instrument manual to determine the
area where the light strikes the cells. In Hach
instruments, it is the bottom of the cell; in
the DRT and Turner, it is the side of the cell.
2a. See section C.I .



la. Preservation of samples is not recommended.
Ib. Within one hour.


TRAINING
GUIDE NOTES

















ifii.B.i
Sp- 21)



-------
WATER MONITORING PROCEDURE:   Determination of Turbidity
 OPERATING  PROCEDURES
        STEP  SEQUENCE
          INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Reagent Preparation

   1.  Turbidi ty Free
      Water
   2. Hydrazine Sulfate
      Solution
1.  Read the turbidity of the
   distilled water.

2.  Filter a quantity of
   distil led water.
3. Read the turbidity of the
   filtered water.

4. If the filtered  distilled
   water shows a lower tur-
   bidity value, treat all
   water used in this pro-
   cedure by filtration.

1. Weigh out 1.00 grams of
   hydrazine sulfate.

2. Transfer the hydrazine
   sulfate to a 100 ml
   volumetric flask.
                           3.
                           4.
   Swirl the flask until
   hydrazine sulfate has
   dissolved.
the
   Dilute with water to the
   100 ml mark.
      la.  Use directions  under "Sample Analysis"  Section.
                                                          2a. Filter using a vacuum membrane filter apparatus
                                                              like that used in the membrane filter technique
                                                              for bacteriological  analysis.
                                                          2b. The membrane filter should have a pore size of
                                                              0.45 micrometer.
      4a.  If the values are the same,  use the distilled
          water without filtration.
      4b.  Check the distilled water  periodically by fil-
          tration to assure absence  of turbidity.


      la.  A trip balance can be used.
      Ib.  Use a plastic weighing boat.

      2a.  Use a plastic wash bottle  and rinse the weighing
          boat with distilled water.
      2b.  Add the washings to the volumetric flask.
      2c.  Rinse three times with about 15 mi's of water.
                                                                                                             E13.C-7

-------
WATER MONITORING PROCEDURE:   WATER MONITORING PROCEDURE
                                                                                                         E13.C-8
OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPE .WING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
  3. Hexamethylenetetra
     mine Solution
  4. Stock Turbidity
     Suspension
     (400 units)
 1. Weigh out 10.00 grams of
   hexamethy1enetetramine.
                          2. Transfer the hexamethy lene-
                             tetramine to a 100 ml
                             volumetric flask.
3. Dilute to the mark with
   distil led water.

1. Pipet 5.0 ml's of the
   hydrazine sulfate solution
   into a 100 ml  volumetric
   flask.

2. Pipet 5.0 ml's of the
   hexamethylenetetramine
   solution into  the same
   100 ml volumetric flask.

3. Mix by swirling the flask.

4. Stopper.

5. Allow to  stand 24 hours.
                          6.  Remove the stopper.

                          7.  After standing,  dilute to
                             the  100 ml  mark.

                          8.  Stopper.
la. A trip balance can be used.
Ib. Use a plastic weighing boat.
Ic. Also called Methenamine.

2a. Use a plastic distilled water wash bottle and
    rinse the weighing boat three times.
2b. Add the washings to the volumetric flask.
2c. Each washing should be about 15 mi's.
la.  Use a 5 ml  volumetric pipet.
Ib.  Always pipet with a pipet bulb.   These chemicals
    are toxic.
                                                         2a. Use a second 5 ml volumetric pipet.
                                                         5a.  The temperature should be between 22°C (72°F) and
                                                             28°C (82°F).
                                                         5b.  The formazine polymer forms  during this time.
                               7a.  With distilled water.

-------
KATER MONITORING PROCEDURE:  Determination of Turbidity
 OPERATING PROCEDURES
        STEP  SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C, Reagent Preparation
   (Continued)
   5. Standard Turbidity
      Suspension
      (40 units)
   6, Standard Turbidity
      Suspension
      (4 units)
9. Mix well.
1. Mix the stock suspension
   wel 1.

2. Pipet 10 ml of the stock
   turbidity suspension into
   a 100 ml volumetric flask.

3, Dilute to the TOO ml mark.

4. Stopper and mix well.
1. Mix the 40 unit standard
   thoroughly.

2. Pipet 10 ml of the 40 unit
   standard into a 100 ml
   volumetric flask.

3. Dilute to the mark.

4, Stopper and mix.
9a. Invert the flask several times while holding the
    stopper in.
9b. This suspension has 400 turbidity units.
9c. This suspension can be kept (when stoppered) for
    one month,

la. This material will  settle out rapidly.
                                                          2a.  Use a 10 ml  volumetric pipet.
                                                          2b.  Pipet with a pipet bulb.
3a. With distilled water,

4a. This suspension is defined as 40 turbidity units.
4b. This suspension can be kept (when stoppered) for
    one week.
4c. Never pour the suspension back into the flask
    after use.

la. By inverting the stoppered flask.
                                                          2a.  Use a  10 ml  volumetric pipet.
                                                          2b.  Use a  pipet  bulb.
                                                          3a.  With  distilled water.

                                                          4a.  This  suspension contains  4  turbidity  units.
                                                          4b.  This  suspension can  be  kept (when  stoppered)  for
                                                              one week.
                                                          4c.  Never pour the suspension back  into  the  flask
                                                              after use.
                                                                                                            E13.C-9

-------
 WATER .MONITORING PROCEDURE:  Determination of Turbidity
                                                                                                         E13.C-1Q
 OPERATING f'FCCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
C. Reagent Preparation
   (Continued)

   7. Standard Turbidity
      Suspension
      (0.4 units)
D.  Instrument
   Calibration
 5.  Label  the  volumetric  flask.


 1.  Mix  the 4  unit  standard
    thoroughly.

 2.  Pipet  10 ml  of  the 4  unit
    standard into a  100 ml
    volumetric flask.

 3.  Dilute to  the mark.

 4.  Stopper and mix.
la. By inverting the stoppered flask.
                                                          2a. Use a 10 ml volumetric pipet.
                                                          2b. Use a pipet bulb.
                                                          3a. With distilled water.

                                                          4a. This suspension contains 0.4 turbidity units.
                                                          4b. This suspension can be kept (when stoppered) for
                                                              one week.
                                                          4c. Never pour the suspension back into the flask
                                                              after use.
                           5.  Label the volumetric flask.
1. Check the meter needle to
   see that it is on the
   zero mark.

2. If it is not, zero it by
   turning the small screw
   located on the meter
   frame.

3. Turn on the power switch.
la.  This is done before the instrument is turned on.
V.D.I
(p.  20)
                                                          3a.  This will  sometimes be a separate switch or the
                                                              instrument is turned on by moving the range
                                                              selector switch to one of the ranges.

-------
HATER MONITORING PROCEDURE:  Determination of Turbidity
 OPERATING  PROCEDURES
        STEP  SEQUENCE
     IMFCRMATION/OPERATING GOALS/SPECiKICATIONS
 TRAINING
GUIDE NOTES
D.  Instrument
   Calibration
   (Continued)
4.  On battery powered in-
   struments, check the
   batteries.

5.  Allow sufficient warm-up
   time.
                           6.  Set the range selector
                              on the 0-1  scale.

                           7.  Shake the 0.4 unit
                              standard suspension.

                           8.  Wait until  large air
                              bubbles disappear.

                           9.  Pour into cell.
                          10,  Wipe the cell  sides  and
                              bottom with a  soft tissue.

                          11.  Insert the cell  into cell
                              compartment.

                          12.  Cover cell  compartment.
4a. All battery powered instruments have a battery
    check position.
4b. See the manufacturer's manual.

5a. This time will vary with the type instrument.
    Follow the manufacturer's manual.
5b. Line operated instruments should be left on if
    they are to be used regularly.
                               7a.  Leave the stopper in place and invert gently
                                   several  times.

                               8a.  Oo not wait too long or suspension will  settle,
                                   usually only several seconds.

                               9a.  With the Hach instruments the  volume must be 25  ml
                                   + 1  ml,   On other instruments  the volume is  not
                                   as critical so  long as the cell  is filled to about
                                   3/4  of its total  volume.
                               9b.  Take care when  handling the cell.   Do not touch
                                   the  bottoms of  the Hach cells  or the lower half
                                   of other type cells.  Handle all  cells by holding
                                   the  top edges.

                              Ida.  Take care that  it is not scratched.
                               la.  Handle near the top section only.
                                                       V.D.10
                                                       (p.  20)
                                                                                                             E13.C-11

-------
                                                                                                         E13.C-12
 WATER MONITORING PROCEDURE:  Determination of Turbidity
 OPERATING  PROCEDURES
         STEP SEQUENCE
      INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Instrument
   Calibration
   (Continued)
13.  Allow instrument  to
    stabilize.
                          14.  Turn  reference  adjust  knob
                              until  the  0.4 mark  is
                              reached.

                          15.  Turn  the range  adjust
                              switch  to  the 0-100
                              scale.

                          16.  Uncover cell compartment.

                          17.  Remove  standard suspension
                          18. Wipe off sides and bottom
                             of  the manufacturer's
                             standard with a soft
                             tissue.

                          19. Insert the manufacturer's
                             reference suspension.
                         20. Cover the cell compartment

                         21. Turn range adjust switch
                             to 0-1 scale.

                         22. Allow the instrument to
                             stabilize.

                         23. Read the turbidity.
13a. Usually in less than one minute.
13b. Stabilization is attained when the needle no
     longer drifts.

I4a. The meter faces vary in how they are marked off.
     The user will have to determine which mark repre-
     sents 0.4 units.

15a. This step will keep the meter needle from bouncing
     off the ends of the meter scale.
15b. May be termed the xlO scale.
                               17a.  Handle near  the  top of  the container.
                               17b.  Retain this  standard  for future use.
                              19a.  If more than one, use the suspension that  is
                                    nearest to 0.4 units.
                              19b.  Handle near the top section only.
                                                        V.D.19a
                                                        (p. 20)
                              23a. If a discrepancy with the expected reading exists,
                                   take note of this reading and proceed.

-------
U'ATER MONITORING PROCEDURE: Determination of Turbidity
 OPERATING  PRCCLDURES
              SEQUENCE
 INFGRMATION/CPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Instrument
   Calibration
   (Continued)
24. Turn the range switch to
    the 0-100 scale.

25. Uncover cell compartment.

26, Remove manufacturer's
    standard,

27. Shake formazine standard.

28. Wipe with tissue.

29, Insert into cell
    compartment.

30. Cover compartment,

31. Allow instrument to
    stabilize.

32, Turn range  switch  to 0-1
    scale.

33, Read the turbidity.
                                                         27a,  The standard used in step 11.
                                                         63b.
If it still reads 0.4 units, there is a dis-
crepancy between the formazine and the manufac-
turer's standard.  Note should be taken of how
much and whenever the manufacturer's standard is
used to calibrate the instrument, this discrepancy
should be added or subtracted from its value.
If it no longer reads 0.4, use the reference ad-
just knob and recalibrate by turning the reference
adjust knob until the 0.4 mark is reached.  Then
repeat these steps beginning at step 15.
Rinse the cell with the next material to be read.
When finished, rinse with turbidity free water.
Store as manufacturer suggests.
                                                                                                             E13.C-13

-------
 HATER MONITORING PROCEDURE;  Determination of Turbidity
                                                                                                         L13.C-14
 OPERATING PROCEDURES
                                     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D. Instrument
   Calibration
   (Continued)
E.  Sample Analysis
34, The formazine standards
    should be used at least
    once/week to monitor
    manufacturer's standard
    for change.

35. For future calibration of
    the 0-1  scale, the manu-
    facturer's standard can be
    used.
                                                         35a.  Taking care to allow for any discrepancy noted
                                                              in step 33.
                                                         35b.  All  ranges on the instrument must be calibrated
                                                              in this same manner.
 1.  Turn on  the instrument.

 2.  Allow warm up time.

 3.  Check batteries.

 4.  Fill  a cell  with  the
    sample.

 5.  Wipe side  and bottom  of
    cell.

 6.  Allow bubbles  to  disperse.
                           7.  Set  the  range  selector
                              on 0-100 scale.

                           8.  Insert cell into cell
                              compartment.

                           9.  Cover the cell compartment
                                                          3a.  If the instrument is  battery operated.

                                                          4a.  Fill  about  three  fourths  of the capacity.
                                                          6a.  A  light  tapping  with  a  finger  will  speed  up
                                                              this  procedure.

-------
UATIR MONITORING PROCEDURE:  Determination of Turbidity
 OPERATING  PROCEDURES
        STEP SEQUENCE
 INrtRMATinN/OFEPATIKG GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E,  Sample Analysis
   (Continued)
10.  Select the range that will
    keep the sample's  reading
    on scale and read  the
    turbidity.

11.  Note which scale is  used.
                                                         lla.

                                                          Ib.
                          12.  Set range  selector on  0-10C
                              range.

                          13.  Uncover cell  compartment.

                          14,  Remove  sample.

                          15.  Select  a manufacturer's
                              standard.
                          16.  Wipe  off  side  and  bottom.

                          17.  Insert  into  cell
                              compartment.

                          18.  Cover cell.

                          19,  Allow instrument  to
                              stabilize.

                          20.  Select  range used  in
                              step  10,
If the value is greater than 40 units, the sample
must be diluted before it can be read.
All samples below 40 units should be read on the
scale that gives the greatest movement of the
meter needle and remains on scale.
                                5a.
Depending on the manufacturer, the standard could
be a simple standard or a set of standards.
Choose the standard that will be in the same range
as the sample.
                                                                                                             E13.C-15

-------
MATER MONITORING PROLLDiM,:  Determination of Turbidity
                                                                                                         E13.C-16
 OPERATING  r
E.  Sample Analysis
   (Continued)
 HI,  Adjust  reference  adjust
     knob  until  standard's
     value is  obtained.

 2?.  Turn  range  selector  to
     0-100 scale.

 23.  Uncover cell compartment.

 24.  Remove standard.

 25.  Insert the cell containing
     the sample.


 26.  Cover the cell compartment

 27.  Set scale on range used in
     step  10 and 20.

 28. Allow instrument to
    stabilize.

 29. Read turbidity,

30. Repeat steps 15 through ?0,
                         31. Turn range selector to
                             0-100 scale.

                         32. Uncover cell compartment.

                         33. Remove eel 1.
                                                    KATri:, GOALS/SPECIFICATIONS
21a. Make any change of standard concentration found
     necessary in step D.33.
                                                         25a.  Care  should  he  taken  not  to  touch  the  cell  except
                                                              at  the  top.   If it  has  been  touched  anywhere
                                                              else, wipe the  cell off.
                                                        30a. If the instrument has drifted, from the value set
                                                             in step 21, recalibrate by adjusting the reference
                                                             adjust knob and reread the samples turbidity,
                                                             beginning at step 21.
                                                          TRAINING
                                                         GUIDE NOTES

-------
HATER MONITORING PROCEDURE;   Determination of Turbidity
 OPERATING PROCEDURES
         STEP SLQttNCL
      INFQSMATION/OPiRATI !\G GOALS/SPEC IFI CAT IONS
                                                     TRAINING
                                                    GUIDE NOTES
E.  Sample Analysis
   (Continued)
F,  Calculations
34,  Insert next sample or if
    no more samples are to be
    read, continue on,

35.  Cover compartment.

36.  Turn off instrument.
                          37.  Wash out cells with
                              turbidity free water.

                          38.  Dry cells with soft
                              tissue.

                          39.  Store cells as manufactur-
                              er recommends.
34a.
                                                         36a,
 1.  Calculate the turbidity
    of the sample.
 la,

 Ib,
If more samples are to be analyzed, repeat
steps 4-10, checking for instrument drift by
checking standardization occasionally.
     If line operated,  leave power on if instrument
     is to be used within a reasonable time.
                                                              Calculation  is  necessary only where the sample
                                                              was  above 40 units  and  had to be diluted.
                                                              If no  dilution  was  performed, then the  turbidity
                                                              of the sample is  read directly from the face  of
                                                              the  meter.
                                                                                        IX.F.Zc
                                                                                        (p. 22)
                                                                                                              E13.C-17

-------
MATER •••QKITQ_R!N_G PROCEDURE:   Determination of Turbidity
                                                                                                          E13.C-18
 OPERATING  PhCCLL-URES
        S1EP  SCQUFNCt
                          GCALS/SPLCIFICATIONS
 TRAINING
GUIDL NOTES
F.  Calculations
   (Continued)
2.  Multiply scale reading
   times the dilution
   factor.
                           3.  Report  the  results.
2a. That is A x (B+C)
               C

    Where B = volume of turbidity free water used
              to dilute the sample (mi's)

          C = the volume of sample used (mi's)

          A = the turbidity units of the dilution
              read on the meter scale
2b, Example:

    If 2 rnl's of sample were diluted to 10 mi's with
    turbidity free water and the dilution had a
    scale reading of 30 units, then

          A = 30
          B = 8
          C = 2
                                                                  x  30 =  150 turbidity units (TU's)
                                3a.  Report as  follows:

                                    TURBIDITY  RANGE
                                       NTU or  TU
                                          0-1.0
                                          1-10
                                         10-40
                                         40-100
                                        100-400
                                        400-1000
                                           >1000
                                                                                            RLCORD TO THE
                                                                                          NEARESTNTU or TU
                                                                                                0.05
                                                                                                0.1
                                                                                                1
                                                                                                5
                                                                                               10
                                                                                               50
                                                                                              100

-------
 WATER MONITORING PROCEDURE:  Determination of Turbidity


                                TRAINING GUIDE


 SECTION                            TQPJX

   I                      Introduction

  II                      Educational Concepts - Mathematics

 III                      Educational Concepts - Science

  IV                      Educational Concepts - Communications

   V*                     Field and Laboratory Equipment

  VI                      Field and Laboratory Reagents

 VII*                     Field and Laboratory Analysis

VIII                      Safety

  IX*                     Records and Reports
 *Training Guide materials are presented here under the headings marked*.
 These standardized headings are used through this series of procedures.
                                                                         E13.C-19

-------
WATER MONITORING PROCEDURES:   Determination  of  Turbidity
 FIELD  AND  LABORATORY  EQUIPMENT
                                                          Section V
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
 D.I
D.10
D.19a
Calibration of any nephelometric turbidimeter should
be done using the standard turbidity suspensions
prepared as described in the Reagent Preparation
Section.  Many manufacturers provide secondary
standards with purchase of these instruments.  It is
well to check this by comparing it to a dilution of
the standard turbidity suspension.

Since the maximum contaminant level is usually one
turbidity unit, dilutions of the standard turbidity
suspension should be made and calibration of the
instrument carried out on the scale near this value.
However, each scale of the turbidimeter and each
secondary standard should be calibrated against the
standard suspension (formazine).

The sample cells to be used with the instruments
must be of clear colorless glass.   They should be
kept scrupulously clean both inside and out, and
discarded when they become scratched or etched.
Some manufacterers claim scratched cells are not a
problem, however, this should be verified.  They
must not be handled at all where the light strikes
them, and should be long enough so that they may be
handled by the top.

This procedure will calibrate the 0-1  scale and
those standards provided by the manufacturer that
can be read on that scale.  If other scales are to
be used, insert a standard suspension  (fonnazine)
that was made up in the Reagent Preparation Section
that can be read on the scale of interest and set
the reference adjust knob to the value.
  El 3.C-20

-------
WATER MONITORING PROCEDURES:   Determination of Turbidity
FIELD AND LABORATORY ANALYSIS
                                                                           Section VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
B.I
Turbidity is one of the parameters classed as a
health limit in the Interim Primary Drinking Water
Regulations.  Public water systems are required to
measure for turbidity daily.  The consents to the
Interim Primary Regulations (p. 59581) indicate
that turbidity measurements were intended to be
carried out by the operators of the public water
system.  This intent was due to the fact that
preservation of the sample was not possible.

The location at which  the sample is to be taken is
listed in the Interim  Primary Regulations as  the
point of entry to the  distribution system.
                                                                                E13.C-21

-------
WATER MONITORING PROCEDURES:   Determination of Turbidity
RECORDS AND REPORTS
                                                          Section IX
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
F,2c
The faces of the meters are labeled as NTU, TU and
FTU.  These are:

   NTU - Nephelometric Turbidity Units
    TU - Turbidity Units
   FTU - Formazine Turbidity Units

These units are considered as being the same,
Express all values obtained as TU's for drinking
waters and as NTU's for wastewaters.
   E13.C-22

-------
                             A PROTOTYPE FOR DEVELOPMENT OF

                             ROUTINE ANALYTICAL PROCEDURES

                                         for the

                   DETERMINATION OF CHLORINATED HYDROCARBON PESTICIDES
                                       as applied in

                            POTABLE WATER TREATMENT FACILITIES
                 National Training and Operational Technology Center
                          Office  of Water  Program Operations
                         U.S.  Environmental  Protection  Agency
CH.PES.lab.WMP.l .11 .77


                                                                              E20-1

-------
WATER MONITORING PROCEDURE:  Determination of Chlorinated Hydrocarbon Pesticides


Operational Procedures

A. Glassware

B, Reagent Preparation

C. Instrument Set-up

   1. Gas Connection
   2, Detector Installation
   3. Leak Check
   4. Column Conditioning

D. Instrument Calibration

E. Standardization of the Sodium Hydroxide

F. Florisil Preparation

   1. Laurie Acid Value Determination
   2. Testing for Proper Elution Pattern

G. Sample Extraction

   1. Pretreatment
   2. Extraction
   3. Concentration

H. Sample Clean-up

I. Sample Analysis

J. Calculations
                                                                        E20-3

-------
    WATER MONITORING PROCEDURE:  Determination of Chlorinated Hydrocarbon Pesticides


    1. Objective:

       To determine the concentration of the chlorinated hydrocarbons, as listed
       in the Interim Primary Drinking Water Regulations, in water samples.

    2. Description of the Analysis:

       A measured water sample is extracted with an organic solvent.   This
       chlorinated hydrocarbon containing solvent is then concentrated to a
       volume of 1 ml.  A small volume (10 yl) is injected into a gas chromato-
       graph and the amount of chlorinated hydrocarbon present is quantitated.
       Should there be interferences presents a method is given to separate  the
       chlorinated hydrocarbon from the interferences.  The equations for cal-
       culating the pi/liter concentration of chlorinated hydrocarbon are given.

       This method is recommended for use only be experienced pesticide analysts
       or under the close supervision of such qualified persons.

       The person attempting to use this outline should have a basic  knowledge  of
       gas chromatography.  Among these skills should be

       a. proper injection technique

       b. proper quantisation technique of the peaks

       c. knowledge of retention times and relative retention times

       d. basic knowledge of the theory and operation of a gas chromatograph

       e. basic chemical skills, such as pipetting, solution preparation, etc.

    3. Applicability of this Procedure:

       a. Range of Concentration

          Many of the chlorinated hydrocarbons can be detected at .001 nig/liter
          quantities.  Of those listed in the Interim Primary Drinking Water
          Regulations, Lindane, Endrin  and Methoxychlor can be determined at this
          level.  Using the concentration inherent in the procedure this level  can
          be lowered.  Toxaphene, also listed in the Interim Primary  Regulations, can
          be determined but is somewhat more difficult.

       b. Pretreatment of Samples

          Upon collection of the sample temperature should be lowered to 4-°  C with
          ice and maintained at this temperature until analyzed.  The maximum holding
          time is 14 days.
E20-4

-------
WATER MONITORING PROCEDURE:  Determination of Chlorinated Hydrocarbon Pesticides


3, Applicability of this Procedure (Continued)

   c. Treatment of Interferences

      This outline includes a clean-up procedure involving separation on a
      Florisil Column,  Inclusion of method blanks during all runs will
      indicate interferences due to impure solvents and reagents.
Source of Procedure:   "Method for Organochlorine Pesticides  in Industrial
Effluents, EMSL, National  Environmental  Research Center,  Cincinnati,  Ohio
45268."
                                                                        E20-5

-------
                       Detm .
                       Det. Limit
                       Extract
                       Sample
                       Inject
                       Sample
                       Compare with
                       Std, Curves
                       P re pa re
                       Elution Column
                       Detm. Laurie
                       Acid Value
                       Elute
                       Sample
                       Inject
                       Eluates
                       Calc.
                       Cone.
If no interf.
calc. cone.
E20-6

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated  Hydrocarbon Pesticides


Equipment and Supply Requirements

A. Capital Equipment:

   1. Gas Chromatograph equipped with

      a.  Glass lined injection port
      b.  Electron capture detector -  tritium or  nickel  63
      c.  Recorder - potentiometric strip  chart (10  in  -  25 cm)  compatible
         with the detector

   2. Gas Chromatographic Column (best  purchased from  gas  chromatographic supply house)

      a.  Tubinq - Pvrex (180  cm lonq  (6 ft.)  x 4mm  ID)
      b.  Glass Wool -  Silanized
      c.  Solid Support - Gas  Chrom Q  (100 -  120  mesh)
      d.  Liquid Phase  - Expressed as weight  percent  coated on solid support

         1)  OV-1, 3%
         2)  OV-210  5%
         3)  OV-17/1.5% plus  QF-1, 1.95%
         4)  QF-1, 6% plus SE-30,  41

   3. Hot Water Bath - Capable of keeping temperature at 50° -  100° C

   4. Source of high quality  distilled  water

   5. Rotometers  - If  the instrument is not  equipped with  meters to monitor
      the flows of gases, these should  be purchased  as options

   6. Analytical  Balance - With a 0.1 milligram  sensitivity

   7. Trip or Platform Balance -  With a 0.1 or 0.01  gram sensitivity

   8. Oven - Capable of maintaining 130°  C

   9, Stop Watch  - Capable of measuring at least  1/2 hour,  the  60 second cycle
      divided to  1/5 second

  10. Cylinder of Argon-methane (95 + 5%)  for  use with pulsed mode detector OR
      Nitrogen -  Purified grade,  moisture and  oxygen free,  for  use with a DC
      mode detector

  11. Pressure Regulator - Two stage with  a  CGA  580  fitting for Nitrogen or a CGA
      350 fitting for  Argon-methane

  12. Filter - For carrier gas  -  molecular sieve  type

  13. Micro  Syringes -  5,  10,  50  ul sizes

  14. pH  Meter -  With  pH electrode
                                                                       E20-7

-------
     WATER MONITORING PROCEDURE:   Determination  of Chlorinated  Hydrocarbon  Pesticides
     Equipment and Supply  Requirements  (Continued)

       15.  Magnetic Stirrer with Teflon coated bar

       16.  Oven (optional) - Forced  air capable of  heating  to  400°  C

       17.  Trap - for oxygen
     B.  Reusable Supplies:
        1.
        2.
        3.
        4.
        5.
       10.
       11
       12,
       13.
Beaker, 150 ml - One for
Beaker, 500 ml - One for
Buret, 10 ml graduations
Buret, 25 ml graduations
each sample - duplicate
each sample - duplicate
- One
- One
                                                      blank
                                                      blank
   Chromatographic Column - Chronaflex (400 mm long x 19 mm ID)  vnth  coarse
   fritted plate on the bottom and Teflon stopcock and a 250 ml  reservoir
   bulb at the top of the column with a flared out funnel  shape  at the  top
   of the bulb, (special  order Kontes Glass -  K-420540-9011) - One for  each
   sample - duplicate - blank
6. Chromatographic Column - Pyrex (about 400 mm long x 20 mm ID) with coarse
   fritted plate on bottom - One for each sample - duplicate - blank
7. Cylinders, graduated
   10 ml - One
   50 ml - Three
   100 ml - Two
   250 ml - Two
   1000 ml - One
8. Dropper Bottle, with dropper, 75 ml  - One
9. Flasks, Erlenmeyer
   25 ml, glass stoppered
   125 ml, glass stoppered
           glass stoppered
           glass stoppered
          , glass stoppered
           volumetric
           Five
            Three
                                    One
                                   - Two
                                   - One
                                   - One
                                    - One
for each sample -  duplicate  -  blank
250 ml,
500 ml,
1000 ml,
Flasks,
10 ml  -
100 ml -
500 ml - Three
Funnels, Separatory, with Teflon stopcock,  200 ml  - One for each  sample  -
duplicate - blank
Glassware,  Kuderna-Danish (K-D), order
a. Concentrator Tube, 10 ml  calibrated,
   Size 1025
b. Snyder Column, three ball,  150 mm long,  #K503000, Size 121
c. Snyder Column, one ball,  150 mm long,  #K5690Q1,  Size 1/19
d. Flask, 500 ml  volume, #K570001
e. Stoppers for flask, f Size  19/22, #K850500
   One for each sample - duplicate - blank
Pipets, graduated
1  ml - Two
10 nl  - One
                                           from Kontes  Glass  Company
                                           f joint 12/22 female,  IK570050,
E20-8

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated Hydrocarbon  Pesticides


Equipment and Supply Requirements (Continued)

  14. Pipets, volumetric
      1 ml - One
      5 ml - One
      10 ml - Two
      20 ml - One
  15, Reagent Bottles,  glass, glass stoppered
      100 ml - One
      500 ml - Four
      1000 ml - One
  16, Ring stand with ring and clamp and  base -  One  for  each  sample  -  duplicate -
      blank
  17. Ruler,  divided in millimeters,  about 30 cm long
  18. Stirring Rod, glass, 12 in, long
  19. Safety Glasses
  20. Bottle, glass, wide-mouth,  with  glass stopper,  500 ml -  One
  21   Desiccator

C. Consumable Supplies;

   1. Alcohol, ethyl, U.S.P.  or absolute,  neutralized to phenolphthalein  (see
      Reagent Preparation Section)
   2. Ethyl ether,  Nanograde, pesticide quality
   3. Florisil,  PR Grade (60-100  mesh), purchase  activated at  1250°  F.  Store
      in the dark in glass containers
   4. Hexane, Nanograde, distilled  in  glass
   5. Laurie Acid,  purified,  CP
   6. Methylene Chloride, Nanograde, distilled in glass
   7. Pesticide  Standards, reference grade
   8. Petroleum Ether,  (boiling range  30-60°  C)  Nanograde  (98+ % pure)
   9. Phenolphthalein Indicator
  10, Soap Solution, any liquid soap mixed 1:1 with water
  11. Sodium Hydroxide,  ACS
  12. Sodium Sulfate, ACS, Granular, anhydrous
  13. Sulfuric Acid, ACS
  14. Distilled  Water
  15. Weighing Boats, plastic disposable
  16. Chart Paper,  for  the recorder
  17. Notebook,  bound
  18. Paper, graph
                                                                       E20-9

-------
 WATER MONITORING PROCEDURE:   Determination of Chlorinated Hydrocarbon  Pesticides
                                                                                                          E20-10
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Glassware Preparation
B. Reagent Preparation
   1.  Ethyl  Ether (6%)
      in Petrn]ptm
1. Clean all Glassware.

2. Wash with soap and water.

3. Rinse with tap water.

4. Rinse with distilled water

5. Muffle at 400° C for 15 to
   30 m1n.
                           6.  Cool  to room temperature.

                           7.  Store until  used.
                                                           3a.  At least  10  times,

                                                           4a.  At least  10  times.

                                                           5a.  Volumetric glassware  should  not be  muffled.
                                                           5b,  Plastic ware and  cap  liners  for sample containers
                                                               should not be muffled.
                                                           5c.  The glassware may be  rinsed  with  redistilled
                                                               acetone followed  by a rinse  with  pesticide
                                                               quality hexane in place  of the muffling.
                                7a. Store inverted or cover mouth with aluminum foil.
                                7b. Sample containers should be stored capped.
1. Use caution with this
   solvent.

2. Add about 100 ml of
   petroleum ether to a 250
   ml graduated cylinder.

3. Pipet 12 ml of ethyl ether
   into the graduated
   cylinder.

4, Dilute to 200 ml with
   petroleum ether.
la.  Prepare and use in a well  ventilated area.
Ib.  Prepare just before use.
                                                           4a.  In the graduated cylinder.

-------
MATER MONITORING PROCEDURE:   Determination of Chlorinated Hydrocarbon Pesticides
OPERATING PROCEDURE:
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                        TRAINING
                       GUIDE  NOTES
  Reagent Preparation
  (Continued)

  2, Ethyl Ether (151)
     in Petroleum Ether
  3. Alcohol - Ethyl
     Neutralized to
     Phenolphthalein
  4. Phenolphthalein
     Indicator
5. Mix with a long glass
   stirring rod.

1. Use the above directions
   except in Step 3, pipet
   30 ml  of ethyl ether.
1. Measure 500 ml  of alcohol
   into a 1  liter Erlenmeyer
   flask.

2. Add 3 drops of phenol-
   phthalein indicator.

3, Titrate with 0.05 N Sodium
   Hydroxide until a pale red
   color is obtained.

4. Store in glass stoppered
   reagent bottle.

1. Weigh out 1,0 g of the
   Phenolphthalein indicator.

2. Transfer to a 100 ml
   volumetric flask,

3. Dissolve in about 50 ml
   ethyl alcohol.

4. Dilute to the mark with
   distilled water.
5a.  Do not store this solution.
    immediately.
Prepare and use
la.  For water supply monitoring for the pesticides
    listed in the Interim Primary Regulations  only
    the 6^ and 15% should be needed.   However, if
    wastewaters are being monitored a 50% eluate will
    have to be prepared and used.
Ib.  Do not store this solution.
la. Use a trip balance.
                                                                                                            E20-11

-------
 WATLR MONITORING  PROCEDtR!"-   Determination of Chlorinated Hydrocarbon Pesticides
                                                                                                          L20-i2
 OPERATING PROCEDURES
                                    INFCRMATiON/OPtkAI J.V3 GQALi/SPLOU ICATIONS
                                                         TRAINING
                                                       GUIDE  NOTES
B,  Reagent Preparation
   (Continued)
   5.  Laurie Acid
      Solution (2% W:V)
   6.  Methylene  Chloride
      (IBS)  in Hexane
      (V:V)
1. Weigh out 10.000 g of
   lauric acid.

2. Add about 250 ml hexane to
   a 500 ml volumetric flask.

3. Transfer the lauric acid
   into the 500 ml volumetric
   flask.

4. Wash the weighing container
   with several small portions
   of hexane and add to the
   volumetric flask.

5. Dissolve the lauric acid.

6. Dilute to the mark with
   hexane.

7. Mix thoroughly.

8. Store in a glass stoppered
   bottle.

1. Add about 200 ml hexane
   to a 500 ml  graduated
   cyl inder.

2. Measure  75 ml of methylene
   chloride.

3, Add the  methylene chloride
   to the hexane.
la.  Use an analytical  balance.
                                                           2a.  Just add hexane to the volumetric flask until
                                                               half full.
                                                           6a.  1  ml  =20 nig lauric  acid.
                                                           2a.  With  a  graduated cylinder.

-------
WATER MONITORING PROCEDURE:  Determination of Chlorinated Hydrocarbon Pesticides
 OPERATING  PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
!. Reagent Preparation
  (Continued)
  7. SuIfuric Acid
     (HS0) 1:1  (V:V)
  8. Sodium Hydroxide
     (NaOH) 10 N
4. Dilute to the mark with
   hexane.

5. Mix.

6, Store in a glass, qlass
   stoppered bottle.
1. Add about 25 ml  of
   distilled water to a 100
   ml volumetric flask.

2, Measure 50 ml concentrated
   sulfuric acid in a 100 ml
   graduated cylinder.

3. Add the sulfuric add to
   the volumetric flask.

4. Dilute to the mark.

5. Cool.

6. Check  to assure the volume
   is still on the mark.

7, Store  in a glass stoppered
   bottle.

1. Weigh  out 40 g of Sodium
   Hydroxide.

2, Measure 100 ml of water.
                                                          2a. Caution:  Use safety glasses.
                                                          2b, Do not add water to acid; follow the procedure.
                                                          3a. The solution will get hot; cool to room tempera-
                                                              ture before proceeding.

                                                          4a. With distilled water.
                                                          6a. If not, add water to mark.
la.  Use a trip balance.
Ib.  Weigh out in a 250 ml  Erlenmeyer flask.

2a.  In a graduated cylinder.
                                                                                                            E20-13

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated Hydrocarbon Pesticides
                                                                                                          E20-14
OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
i. Reagent Preparation
  (Continued)
  9. Sodium Hydroxide
     (NaQH) 1.0 N
 10.  Sodium Hydroxide
     (NaOH)  (0.05  N)
 3. Add  the water to the
   Erlenmeyer  flask.

 4. Swirl  to dissolve.
 5. Stopper the flask with
   a rubber stopper and
   label.

 1, Weigh out 20 g of sodium
   hydroxide,

 2. Transfer the NaOH to a 500
   ml Erlenmeyer flask.

 3. Measure 500 ml  of water,

 4. Add the water to the
   Erlenmeyer flask.
                          5.  Swirl  to dissolve.
6. Stopper the flask with a
   rubber stopper and label.

1. Add about 250 ml  distilled
   water to a 500 ml
   volumetric flask.

2. Transfer 25 ml of the 1.0 N
   NaOH to the volumetric
   flask.
3a. The solution will get hot; cool under cold water.
3b. Use safety glasses.

4a. Do not stopper and invert to mix.
4b. Use a stirring rod or magnetic stirrer and
    Teflon coated bar.

5a. Do not stopper unless cool,
la. Use a trip balance.
Ib. Weigh out in a plastic weighing boat.
3a. Us? a 500 ml qraduated cylinder.

4a. The solution will get hot; cool  under
    cold water.
4b. Use safety glasses.

5a. Do not attempt to stopper and mix by inversion,
    use a stirring rod or a magnetic stirrer and
    Teflon coated bar.

6a. Do not stopper unless cool.
                                                         2a, Use a 25 ml  volumetric pi pet.

-------
 WATER  MONITORING  PROCEDURE;   Determination of Chlorinated Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STEP
    IflFCRK-'JlON/OPESATING  GOALS/SPECIFICATIONS
                                                    TRAINING
                                                   GUIDE NOTES
8. Reagent Preparation
   (Continued)
  11.  Sodium Sul fate
3. Dilute to the mark,

4. Mix thoroughly.

5. Store in a glass, rubber
   stoppered, bottle.

1. Weigh out 1.0 gram of the
   sodium sulfate.

2. Add enough methylene
   chloride - Hexane
   (Reagent 6) to cover the
   sodium sul fate.

3. Mix.

4. Decant reagent.

5. Inject 10 pi  of the solvent
   into the gas  chromatograph.

6. If contaminated proceed as
   below,

7. Weigh out about 100 g of
   sodium sulfate in a 500 ml
   beaker.

8. Place beaker and sodium
   sulfate in an oven.

9. Heat the sodium sulfate
   for 4 hours at 400° C.
la.
Ib.
Use a trip balance.
Weigh in a 150 ml  beaker.
                                                           4a.

                                                           5a.
    Take care not to include any sodium sulfate.

    After retention times and optimization have been
    determined.
                                                               Contamination  is  shown  peaks  of  1  times  the
                                                               noise  level  are obtained.
6a


7a. Use a trip balance, weigh  in the beaker.
                                                           8a.  The  oven  should  be  preheated  to  400°  C.
                                                                                                              E2Q-15

-------
 WATER  MONITORING  PROCEDURE:   Determination  of  Chlorinated  Hydrocarbon  Pesticides
                                                                                                         E20-16
 OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B. Reagent Preparation
   (Continued)
  12.  Pesticide  Standards
10. ReTOve from oven and place
    in a desiccator and allow
    to cool.

11. Transfer  to a  wide-mouth
    glass stoppered bottle.

12. Store in  the bottle in  a
    desiccator.

 1. Consult Tables 1  and 2
lOa.  Use tongs  to handle the hot beakers.
lOb.  Be sure the desiccant is activated.
 la.  In order to  properly  calibrate  the  instrument
     both  qualitatively  (Retention Values)  and
     quantitatively  (Detection  Limits) single
     standard and  a  mixture  containing all  the
     pesticides of interest  should be made.
 Ib.  The amount weighed  out  to  each  compound was
     kept  at  10 mg in  order  to  have  a weight which
     could be accurately weighed yet using  the
     smallest amount of  compound possible.
 Ic.  After weighing  out 10  mg  and preparing 10 ml of
     the stock solution  of each pesticide,  proceed
     toward the right  of the table and make two
     dilutions as  directed.
 Id.  Use Table 2 for preparing  two mixtures.  Mixture
     1  is  used to  prepare  the column.  Mixture 2
     will  be  used  to determine  the standard curves.
                                                                                                                 I.B.12

-------
                      DIRECTIONS FOR DILUTION TABLE

In order to properly calibrate the instrument both qualitatively (Retention
Values) and quantitatively (Detection Limits) a single standards and a mixture
containing all the pesticides of interest should be made.

The amounts weighed  out of each compound was kept at 100  mg in order to have
a weight which could be accurately weighed yet using the smallest amount of
compound possible.

After weighing out and diluting the stock solution of each pesticide, proceed
toward the right of the chart.  Making dilutions as directed.

Use Table 2 for preparing two mixtures.   Mixture 1  is used to  prepare the
column.  Mixture 2 will be used to determine the standard  curves.
                                                                        E20-17

-------
                                               TABLE 1
                                                                                                E20-18
Compound

Lindane

Endrin

Methoxychlor

Aldrin

Toxaphene

Stock

10 mg
10 ml* ;
10 mg
100 ml* 1
10 mg
10 ml** 1
i
1
10 mg
10 ml*
10 mg ;
10 ml* ;
Dilution 1

4 ml stock
100 ml**
2 ml stock
100 ml**
1 ml stock
1 100 ml**
1 ml stock
100 ml**
5 ml stock
1 100 ml**
Cone, mg/1
;
1
40

i
20 |

i
10 j

10

50 j
1
Dilution 2
i

1 ml Oil. 1
100 ml
1 ml Oil. 1
100 ml
None

1 ml Oil. 1
100 ml
1 ml Oil. 1
100 ml
Cone, mg/1

0.4

0.2

None

0.1

0.5
i
 * Make stock solutions in 2.2.4 Trimethyl  pentane (Isooctane)
** Use hexane for all  dilutions of the stock.

-------
TABLE 2
                            Use Hexane for all  dilutions
Compound
Lindane
Endrin
Methoxychlor
Aldrin
Toxaphene
Add to th
Amount ml .
1 .0
1.0
1.0
1.0
1.0
Mixture 1
e same 5 ml vo
Dilution Ho.
Stock
Stock
Stock
Stock
Stock
. flask
Cone, mg/1
200
200
200
200
200






Add ti
Amount ml .
1
1
1
1
1
Mixture 2
3 the same 100 ml vol .
Dilution No.
1 (40 mg/1)
1 (20 mg/1)
Stock (1000 mg/1)
1 (10 mg/1)
1 (50 mg/1)
flask
Cone, mg/1
0.4
0.2
10.0
0.1
0.5
                                                    E20-19

-------
 WATER MONITORING PROCEDURE:   Determination  of Chlorinated  Hydrocarbon  Pesticides
                                                                                                         E20-20
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERAT ING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
C. Instrument Set-up
   1,  Gas Connection
   2.  Detector
      Installation
   3.  Carrier  Gas Leak
      Check
1. Remove cylinder cap from
   cylinder of carrier gas.
                           2.  Install  the  pressure
                              regulator.
                           3.  Connect  the  cylinder  to
                              the  instrument.
1. Install  the electron
   capture  detector.
1. Open master valve on the
   cylinder,

2. Adjust regulator control,
                           3.  Close  carrier  gas control
                              valve  on  the instrument.

                           4.  Turn off  master cylinder
                              va1ve.
la. The tank should be chained to the wall  or lab
    bench.
Ib. Use Argon-methane for pulsed mode detector or
    Nitrogen for detector operated in a DC  mode.
    Consult the manufacturer's manual.

2s. The regulator should have a CGA 350 fitting
    for Argon-methane or for Nitrogen.
2b. Fitting should be tight to a CGA 350 fitting
    to prevent leaks.

3a. Use teflon tape on all metal threads to prevent
    leaks.
3b. Polethylene (1/8" diameter) tubing can  be used.
3c, If plastic tubing is used, nylon ferrules
    should be used with the connector fittings.

la. This can be of the tritium or nickel 63 type.
Ib. See the manufacturer's manual on procedures for
    installation.
                                                           2a. To about 65  psig and allow to stabilize  (about
                                                              1 minute).
V.C.I
(p- 45)

-------
 WATERMONITOR!NG  PROCEDURE.:    Determination  of  Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STLP SEQUENCE
                                    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Instrument Set-up
   (Continued)
   4. Column
      Conditioning
5.  Observe gage on the
   cylinder.
6. Use soap solution and
   check for leaks at the
   injection port and column
   connections.

1. Install  packed column in
   the oven by connecting
   only the column inlet.
                                5a.  Pressure should not drop  more  than  a  few  psig.
                                5b.  If the  pressure does drop use  soap  solution  to
                                    locate  leak.
                                5c.  Correct leak  and check  again.
                           2. A flow of carrier gas
                              should be started through
                              the detector.
                           3. Begin a low flow of
                              carrier gas through the
                              column.
                           4. Wait 5 minutes.
                                la.  These columns  should  be  purchased  and  meet  the
                                    specifications  as  listed under  the equipment
                                    section.
                                Ib.  Column conditioning  is essential to eliminate
                                    column bleed and to  provide  acceptable analysis.
                                Ic.  Do not connect the column to the detector.
                                Id.  If in doubt as  to  column installation  refer to
                                    the manufacturer's manual.

                                2a.  Use the purge  gas  line or in dual  column oven
                                    by connecting  an unpacked column to the detector,
                                2b.  In some systems it may be necessary to
                                    temporarily connect  the  carrier gas to the  air
                                    or hydrogen inlet  in  order to get  a flow to the
                                    detector.   The manufacturer's manual should be
                                    consulted.

                                3a.  Less than  60 ml/min.(N40-50).
                                3b.  Removes oxygen and other trapped gases.
                                3c.  This will  be two separate flows.   The  column
                                    should not be  connected  to the  detector.
                                                                                                             E20-21

-------
 WATER MONITORING PROCEDURE:    Determination of Chlorinated Hydrocarbon  Pesticides
 OPERATING PROCEDURES
STEP SEQUENCE
                                    If,FORMAT10N/CP£?ATING  GOAL^/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
C.  Instrument Set-up
   (Continued)
5. Turn on power to the oven
                           6.  Adjust column  temperature
                              to  near the  maximum  re-
                              commended  temperature  for
                              the liquid phase  being
                              used.
                           7.  Continue  heating  for
                              2  hours,

                           8.  Reduce  temperature  to
                              about 40°  C below
                              maximum temperature.

                           9-  Allow column  temperature  to
                              equi1ibrate.
                          10.  Check  carrier  gas  flow.
                        5a.  Consult  manufacturer's  manual  for  location  and
                            necessary  steps.
6a.
                                  Column
                                                       Max.  Temp.
                                                          0  C
                                OV-17  &  QF-1        250
                                OV-210             275
                                OV-1                350
                                QF-1 & SE-30        250
                       8a.  See  table above  (6a).
                        9a.  Minimum of  3C minutes.
                        9b.  Caution:  Bleed off of  the  liquid phase will
                            occur  if  the column temperature  is not fully
                            equilibrated.

                       lOa.  About  50  ml/nrin.
                       lOb.  If  the instrument does  not  come  equipped with
                            rotometers  to monitor the flow rate of the
                            carrier gas this should be  purchased as an
                            option.
                       lOc.  The pressure regulator  on the cylinder should
                            be  set at 65 psig.
                       lOd.  The electron capture detector must be installed,
                          II. Allow to remain at
                             temperature and flow for
                             one hour.
v.c.e.b
(p.  45}

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
         STEP  SLQUEfiCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C.  Instrument Set-up
   (Continued)
12.  Increase temperature  to
    about 20° C above
    operating temperature.
                          13,  Continue the same  flow  of
                              carrier gas.

                          14.  Allow to equilibrate  for
                              24-48 hours.

                          15.  Turn off oven and  allow
                              to cool  to room
                              temperature.

                          16.  Adjust carrier gas  flow to
                              method flow rate.

                          17.  Connect column to  the
                              detector.

                          18,  Turn on oven  and adjust to
                              method column temperature.

                          19.  Allow the  instrument  to
                              equilibrate at least  one
                              hours.
12a. This is operation temperature not maximum
     temperature.
12b. These temperatures would be
       200° C for  the OV-17 and qF-1  and the
       QF-1 & SE-30
       OR
       200° C for  the OV-210 and the  OV-1.
                                14a.  Caution:   Do  not  exceed  maximum  recommended
                                     temperatures.   See  6a.  this  section.
                                16a.  60  rnl/min.  for  the  OV-17  & QF-1 and QF-1  & SE-30.
                                     70  ml/min.  for  the  OV-210 and OV-1.

                                17a.  Check  connection with  soapy solution.
                                18a.  200°  C  for  the  OV-17  & QF-1 and  the QF-1 & SE-30,
                                     180°  C  for  the  OV-210 and  the OV-1.

                                19a.  Preferably  overnight.
                                                                                                             E20-23

-------
 MATER MONITORING  PROCEDURE:    Determination  of  Chlorinated Hydrocarbon Pesticides
                                                                                                         E20-24
 OPERATING PROCEDURES
        STEP SEQ'JENCt
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Instrument
   Calibration
   1.  Optimization
   2.  Retention  Time
      Determination
1. Check instrument operating
   conditions after the
   instrument has been set-up
                           2.  Inject  5  ul of standard
                              mixture 2.
                           3.  Adjust  the operating
                              parameters to achieve
                              optimum results.
1. Inject 5 yl  of dilution
   2 (Table 1)  of the
   standards one at-a-time.
la. The oven temperature should be stabilized.
Ib. The flow rate constant.
Ic. The column conditioned.
Id. The flow system checked for leaks.

2a. This standard mixture was prepared  from the
    stock standards (Table 2).
2b. The actual  volume of material  injected should be
    kept constant.  That is, the same for standards
    and samples.
2c. The standard mixture 2 can be used  to optimize
    the instrument initially and thereafter monitor
    its performance.

3a. Best resolution (separation of peaks) and
    retention times can be achieved by  adjusting
    the column temperature and/or the carrier gas
    flow rate.
3b. Compare results with the standard chromatograms
    attached.
3c. Caution should be taken to allow the instrument
    to equilibrate after any changes.
3d. Optimum results would include good  separation of
    the peaks,  good sensitivity, good reproducibility,

la. The individual standards lindane, endrin,
    methoxychlor, aldrin, and toxaphene should  be
    used.
Ib. Dilutions of the  individual standards can be
    prepared from the stock solutions as in Table I.
Ic. Injection is a technique which must be learned
    and practiced in  order to make accurate and
    reproducible injections.
Id. The analyst should consult a text on gas
    chromotography or a syringe manufacturer's
    1iterature.
                                                                                       V.D.1.3a
                                                                                       (p.  45)
                                                                                                                 VII.D.4
                                                                                                                 (p.  46)

-------
 WATER  MONITORING  PROCEDURE:   Determination  of Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                    TRAINING
                                                   GUIDE NOTES
D. Instrument
   Calibration
   (Continued)
   3. Detection Limit
      of the Instrument
2. With a stop watch measure
   the time elapsed between
   the first appearance of
   the solvent peak and the
   peak of the known
   standard.

3. Repeat steps 1  and 2 six
   more times.
1 .  Inject 5 ul  of a single
   pesticide standard.
                              Continue with each  standard
                              by diluting the last
                              standard concentration
                              run in half and injecting
                              5 ul  portions unti1  a
                              detection limit is  obtainec

                              The peak height in  milli-
                              meters should be used and
                              plotted against the  known
                              concentrations to produce
                              a standard curve.
2a.
2b.
This time is called the retention time.
After the retention times have been obtained
using single standards, the mixtures should
be used to determine that no changes in  retention
time occur.
                                                           3a.
    At least 7 repeat times should be obtained and
    the mean value obtained for each pesticide.
3b. If in subsequent injections, the retention times
    vary significantly (+ 2%)  the system should  be
    checked over.

la. Begin with Dilution 2,  Table 1 of each.
Ib. The standards need not  be  carried through
    the extraction and concentration steps  to
    develop a standard curve.

2a. The detection limit is  usually considered  as a
    standard whose peak is  2 times the highest peak
    caused by noise and is  run at the most  sensitive
    setting of the instrument.
                                3a.  The curve produced  can  be  used  to  select  the
                                    concentration of standard  to  be injected  to
                                    calculate the concentration of  an  unknown.  See
                                    calculation of results  section.
                                3b.  To select the concentration to  be  used  to
                                    calculate the unknown,  first  note  the peak height
                                    of the unknown.   Second refer to this standard
                                    curve and find a concentration  which  gave the
                                    same peak height and  inject it  into  the
                                    instrument.  This should provide a standard
                                    very close to the sample peak height  to use in
                                    the calculations.
                                                                                       V.D.3.2
                                                                                       (p. 45)
                                                                                                            E20-25

-------
 WATER MONITORING PROCEDURE!:   Determination of Chlorinated Hydrocarbon Pesticides
                                                                                                          E20-26
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE NOTES
E. Standardization of
   Sodium Hydroxide
F.  Florisil  Preparation
   1.  Laurie Acid Value
      Determination
 1. Weigh out TOO mg of  lauric
   acid.

 2. Quantitiatively transfer
   to a 125 ml Erlenrneyer
   flask.

 3. Dissolve with 50 ml
   ethyl alcohol.

 4, Add 3 drops phenolphthaiein

 5. Titrate with  (0.05 M)
   sodium hydroxide.

 6. Calculate mg  lauric  acid/
   ml 0.05 fl NaOH.
la. On an analytical balance.
                                                           3a. Neutralized to phenolphthalein (See B.3).
                                                           5a. See reagent section for preparation.
                                                           5b. Until  a pink color persists.
                                                           6a. ma lauric acid
                                                               ml  of" 0.05 t; JiaOH =
                            100 mg Jaurjc acid
                            number of ml needed
                            to titrate (from Step 5]
                                                           6b.  See  calculation section J.I  for use.
1. Place 2.000 grams of
   Florisil in a 25 ml  glass
   stoppered Erlenmeyer flask.

2. Cover the flask loosely
   with foi1.

3. Heat overnight at 130° C.

4. Stopper and cool to room
   temperature.
la.  Weigh  on an analytical  balance,
VII.F.I
(p.  46)

-------
 WATER MONITORING PROCEDURE:    Determination  of  Chlorinated  Hydrocarbon  Pesticides
 OPLRATiNG PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Florisil  Preparation
   (Continued)
5, Remove stopper and add
   20.0 ml  of lauric acid
   solution (reagent 5).

6. Stopper the flask.

7. Swirl intermittently for
   15 minutes.
                           8. Allow the florisil  to
                              settle.

                           9. Pipet 10.0 ml  of the
                              supernatant into a  125 ml
                              Erlenrneyer flask.
                          1C. Add 5C ml ethyl alcohol.
                          11. Add 3 drops of
                              phenolphthalein indicator,

                          12. Titrate with 0.05 N NaOH
                              to a permanent red color.
                          13. Calculate the amount of
                              lauric acid absorbed by
                              the florisi1.
Sa, Use a 20 ml  volumetric pi pet.
                                                           7a.  A  gentle  swirling  of  the  flask  is  sufficient.
                                                           7b.  This  is  to  assure  contact of  the  florisil with
                                                               the  lauric  acid  solution.
                                9a.  Use a 10 ml  volumeteric pi pet.
                                9b.  Supernatant  is the clear liquid standing above
                                    the settled  absorbent.
                                9c.  Avoid inclusion of any  florisil.

                               lOa.  Neutralized to phenolphthalein.
                               lOb.  See reagent section for preparation.
                               lla. The solution should remain colorless.

                               12a. Those individuals not familiar with procedures
                                    for titration should consult a procedure on the
                                    use of a buret.

                               13a. The calculation will be found under the
                                    calculation section of this procedure.
                               13b. This lauric acid value must be obtained for each
                                    new batch of florisil purchased.  Then an
                                    equivalent weight of the new batch can be
                                    calculated to obtain values similar to the old
                                    batch.
                                                                                                             E20-27

-------
 WATER  MONITQRINS  PROCEDURE:   Determination  of Chlorinated  Hydrocarbon  Pesticides
                                                                                                        £20-28
 OPERATING PROCEDURES
         STEP  SEQUENCE
     INFORMATION/OPE PAT ING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE  NOTES
F.  Florisil  Preparation
   (Continued)

   2.  Testing for Proper
      Elution Pattern
14.  Store the rest  of the
    florisil.

 1.  Weigh out the determined
    amount of florisil.
                           2.  Pour  the  florisil  into  the
                              chroma tographic column.

                           3.  Tap the column lightly  to
                              settle the  florisil and  to
                              level  its surface.

                           4.  Add about one-half  inch
                              of sodium sulfate  to the
                              top of the  florisil in
                              the column.

                           5.  Add 40-50 ml of petroleum
                              ether  to the top of the
                              column.
                          6. Time and rate of passage
                             of the petroleum ether
                             through the column.

                          7. Remove the container used
                             to collect the petroleum
                             ether.
14a.  In glass bottle with glass  stopper kept in  the
     dark in an oven at 130C  C,

 la.  Determined from the 1 auric  acid value  of the
     batch of fIon'si 1.
 Ib.  Usually about 10 to 15 grams.
 Ic.  See the calculation section (J.2).

 2a.  The column should  meet the  specifications listed
     in the equipment section.

 3a.  The column should  not be packed so tight as  to
     impede solvent flow.
                                4a. This should be  the anhydrous, granular, ACS grade
                                4b. Should be pre-conditioned by heating at 400° C
                                    for  four hours  (See B.ll).
                                5a. This is used to pre-wet the column.
                                5b. The column may generate heat as it is wet by the
                                    solvent.  Let cool to room temperature before
                                    proceeding,
                                5c. This solvent can be collected in any type of
                                    container and need not be saved.

                                6a. This elution rate should be set at about 5 ml
                                    per minute.
                                6b. Use a 10 ml graduated cylinder and a stopwatch.

                                7a. This should be done while a small  level of
                                    solution is still  above the sodium sulfate
                                    layer.
VII.F.2
(p. 46)

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STEP  SEQUENCE
     INFORMATION/OPERA:ING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Florisil  Preparation
   (Continued)
8. Place a clean 500 ml  K-D
   flask equipped with a 10
   ml  graduated ampul  under
   the column,

9. Just prior to the exposure
   of  the sulfate layer to
   the air add 10 ml of
   mixture 2.
                          10.  Rinse  container with 10  ml
                              of petroleum ether.

                          11,  Add 200 ml  of the  6« ethyl
                              ether  in petroleum ether
                              solution.
                          12.  Collect the  200  ml
                          13.  Add 200  ml  of the  15%  ethy
                              ether-petroleum ether
                              solution,

                          14.  Immediately replace  the
                              500 ml  K-D  flask with
                              another  clean one.
 8a.  Some portion of the petroleum ether will  be
     collected.
 8b.  K-D is a Kuderna-Oanish.
 9a.  See the reagent preparation section (B.12, Table
     2).
 9b.  The sulfate surface should not be allowed to dry
     between additions of the mixture and the
     following eluates.

lOa.  Add the rinse to the column.
                               lla.  Measure in a graduated cylinder.
                               lib.  If prepared just prior to use by  the directions
                                    given in the reagent preparation  section,  only
                                    200 ml  will  be prepared and can be completely
                                    transferred to the column.
                               lie.  Add small  portion slowly to bring liquid level
                                    to the  top of the florisil  column then add the
                                    rest.

                               12a.  Close the  column's stopcock before the sulfate
                                    layer is exposed to the air.
                                                                                                              E20-29

-------
 WATER MONITORING PROCEDURE:    Determination  of Chlorinated  Hydrocarbon  Pesticides
                                                                                                        [20-30
 OPERATING PROCEDURES
        STEP SEQUENC!
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                     TRAINING
                                                    GUIDE NOTES
F.  Florisil  Preparation
   (Continued)
15.  Collect the 151 eluate.
                           16.  Just  prior  to  the  sulfate
                               layer being  exposed  add
                               200 ml  of ethyl  ether.
                           17.  Immediately  replace  the
                               500  ml  K-D flask with
                               another clean one.

                           18.  Collect the  ethyl ether
                               elution.

                           19.  Insert  a Snyder column
                               into each flask.

                           20.  Concentrate  each elution.
                          21. Remove the flask from the
                              ampul .
                          22. Rinse the walls of the
                              ampul .

                          23. Stopper the ampul.
15a,
                               16a.
There will be a small overlap; in this case a
small portion of the 6% will be collected in the
15% flask.  Again close the stopcock to prevent
exposure of sulfate layer to the air.

For wastewater or if other pesticides are to be
monitored other than those listed in the Interim
Primary Regulations, a 50% ethyl  ether in
petroleum ether elution would be carried out
at this time followed by the straight ethyl
ether elution.
                              203,
                              21a,
                              22a.
                              22b.
     Use the  procedure  under sample  concentration,  G.3,
     1  through  6.
     Rinse  bottom of  flask  and  lower  glass  joint  into
     the  ampul.

     Rinse  to  a  final  volume  of 10  ml.
     Use  a  5.0 cc glass  syringe.

-------
 HATER  MONITORING PROCEDURE:    Determination  of  Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STLP StgbENCE
     HiFORMATIO'.'/OPF.SATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Florisil  Preparation
   (Continued)
24.  Inject 5 pi  of the
    fractions.
                           25.  Compare retention  times
                               with standards  run in
                               Section D.5.
G. Sample Extraction and
   Concentration
   1.  Pre-Treatment
   2. Extraction
 1.  Blend the sample.

 2.  Adjust pH to 6.5 -  7.5
 1. Measure out 1  liter of
    sample.
                            2.  Transfer to a  2 liter
                               separatory funnel.
24a.  The pattern will  be 6K eluate
     Aldrin-Lindane-Methoxychlor-Toxaphene,  15% eluate
     Endrin, Ethyl  ether and anytning  left on the
     column.

25a.  If agreement is not attained, check the instrument
     operational parameters (Section D.I).
 la.  This is usually not required for drinking waters.

 2a.  Use 50?t sulfuric acid or 1.0 N sodijm hydroxide.
 2b.  Use a pH meter to measure the pH.
 2c.  Usually not necessary for drinking waters.

 la.  Use a 1 liter graduated cylinder.
 Ib.  Experience witn tie sample source will  indicate
     if smaller volumes should be used.  If  smaller
     volumes are used they should be diluted to 1
     liter volume before extraction.
 Ic.  A 1 liter volume of distilled water should be
     carried through the entire procedure along with
     each sanple batch to serve as a metnod  blank.
 Id.  The standard mixture 2 to be used as a  check  on
     the curve (D.3.3) must also be carried  through
     this procedure.
 le.  Duplicate analysis of the samples are recommended
     to be carried out.
 If.  Surface waters sometimes require larger volumes.

 2a.  Use a 2 liter size in order to have room to
     obtain proper mixing.
                                                                                                                  VII.G.2.If
                                                                                                                      6)
                                                                                       (P.
                                                                                                              E20-31

-------
 MATER MONITORING PROCEDURE:   Determination  of  Chlorinated  Hydrocarbon  Pesticides
                                                                                                         E20-32
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G. Sample Extraction and
   Concentration
   (Continued)
 3. Add 60 ml of the methylen«
   chloride-hexane mixture
   (Reagent 6) to the
   separatory funnel.

 4. Shake vigorously for two
   minutes.
                            5.  Return  to  holder and allow
                               mixtures  to  separate.
                            6.  Prepare  the anhydrous
                               sodium sulfate column.
7. Remove the stopper from
   the separatory funnel.

8. Drain the water layer into
   a one liter Erlenmeyer
   flask.
3a. Use the 60 ml to rinse the sample container (if
    empty) and the graduated cylinder.
4a. Pressure may build up in the separatory funnel.
    Invert, with stopper tightly in place, open
    stopcock slowly to relieve pressure, do this
    several times during shaking.

5a. The water layer will be on the bottom and the
    methylene chloride - hexane on top.
5b. The holder is some type support, such as a ring
    clamped to s ring stand.

6a. Add 3 to 4 inches of anhydrous sodium sulfate to
    the chromatographic column (Pyrex, approximately
    400 mm long x 20 mm I.D. with a coarse frit on
    the bottom).
6b. The column should be placed in a support such as
    a  clamp attached to a ring stand.
6c. Position the bottom of the column  well into the
    neck of a 500 ml Kuderna-Danish flask with a 10
    ml  graduated ampul  attached to the stand.
6d. The sodium sulfate should have been heated at
    400° C for four hours in an oven (Section  B.ll).
                                                          8a. Keep the water layer for further extractions

-------
 MATER MONITORING PROCEDURE:   Determination  of  Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G. Sample Extraction and
   Concentration
   (Continued)
 9.  Pour the organic
    (methylene  chloride-
    hexane)  layer  into the
    sodium sulfate column.

10,  Collect the organic layer
    after it has passed
    through  the column in  a
    500 ml   Kuderna-Danish
    (K-D) flask with a 10  ml
    ampul attached,

11.  Return the  water phase to
    the separatory funnel,

12.  Rinse the Erlenmeyer  flask
    with a second  60 ml
    portion  of the methylene
    chloride-hexane solvent.

13.  Transfer the second
    portion of the methylene
    chloride-hexane solvent
    to the separatory funnel.

14.  Put stopper in place.

15.  Shake vigorously for  two
    minutes.

16.  Repeat steps 5 and 7
    through  11.
9a. Pour from the top of the separatory funnel.
9b. Use a funnel  at the top of the column to help
    transfer sample into column.
                                                          15a.  Release  pressure  periodically.
                                                          16a.  Step  6  does  not  have  to  be  done  again.  Use the
                                                               same  column.
                                                          16b.  Collect the  second  organic  layer (methylene
                                                               chloride-hexane)  in the  same  K-D flask.
                                                                                                             E20-33

-------
 WATER MONITORING  PROCEDURE:    Determination  of  Chlorinated  Hydrocarbon  Pesticides
                                                                                                        E20-34
 OPERATING PROCEDURES
        STEP SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G.  Sample Extraction and
   Concentration
   (Continued)
   3.  Concentration
17. Perform the extraction
    procedure with a third
    portion of the methylene
    chloride-hexane solution,

18. Rinse the sodium sulfate
    column three times  with
    10 ml  volumes of
    methylene chloride-hexane.

 1. Insert a condenser  into
    the K-D flask.

 2. Place K-D flask in  a
    holder above a boiling
    water bath.

 3. Add small  boiling chip.

 4, Lower the ampul  into  the
    water.
17a. Steps 12 through 17.
17b. All  three 60 ml  organic extraction portions  are
     combined in the K-D flask.
                                                          18a.  Collect  in  the K-D.  flask.
 la.  Condenser - Snyder column - three ball
                                                          4a. The water level should be maintained below the
                                                              lower joint  (where the ampul connects to the
                                                              flask).  The lower rounded surface of the flask
                                                              should be bathed in steam.
                                                          4b. Surrounding  the flask with aluminum foil will
                                                              help.
                                                          4c. The evaporation must not go  to dryness.  The
                                                              analyst should stay with the flask.
                                                Water
                                                                                    - -
                                                                                    Level

-------
 UATER MONITORING PROCEDURE:    Determination of Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
       STEP SEQUENCE
If.'FORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDL NOTES
G. Sample Extraction and
   Concentration
   (Continued)
5.  Allow the solvent to
   evaporate.
                            6.  Concentrate to  about 1  ml
                               volume.
                            7.  Remove from the bath.

                            8.  Remove the  three ball
                               condenser from the  flask.

                            9.  Rinse the lower joint  of
                               the column  into the ampul.

                           10.  Insert a  micro Snyder
                               column (one ball)  into the
                               ampul.

                           11.  Return to the  water bath.
                           12.  Boil  and  reduce  the  volume
                               to about  0.2  to  0.5  ml.
                           13.  Cool .

                           14.  Remove  the  micro  column.
5a. Evaporation should be adjusted so that the
    solvent vapors are rushing through the condenser.
5b. Carry out the evaporation in a hood.
5c. There should be no splashing at the top of the
    column or flooding of the chambers.

6a. This will be in the ampul.  However, after
    cooling solvent that has remained in the Snyder
    column will  drain back into the ampul  and raise
    the volume to above 1 ml.
                               9a.  Rinse  with  hexane.
                              lla.  Caution:   When  using  the  micro  column, the  analyst
                                   should  constantly  watch  the  samples  as  they  are
                                   heated.   They must not go to dryness.

                              12a.  Do  not  allow to  go to dryness.
                              12b.  After cooling the  volume  will  increase  by  about
                                   0.1  ml  by the solvent draining  back  into  the
                                   ampul.
                                                                                                             E20-35

-------
 WATER  MONITORING  PROCEDURE:   Determination  of  Chlorinated Hydrocarbon Pesticides
                                                                                                         E20-36
 OPERATING PROCEDURES
              STEP  SEQUENCE
                                    INFORMATION/OPERATING GOALS/SPECIFICATION!
                                                                                     TRAINING
                                                                                    GUIDE NOTES
G. Sample Extraction
   Concentration
   (Continued)
and
15.
                          16.
Rinse the lower joint of
the column into the  ampul.
         Dilute  to
         of 1  ml.
              the  final  volume
                          17.  Stopper  to prevent further
                              evaporation.

                          18.  Inject 5 ul of this
                              concentration into the gas
                              chromatograph.
                          19.  If interferences are
                              present proceed to the
                              clean-up section.

                          20.  If no interferences are
                              present and pesticides are
                              determined, proceed to the
                              calculation section.
15a.  Use a 10 ul  syringe.
15b,  The rinse should be about 0.1  to D.2  ml.
15c.  Caution:  Do not go above 1  ml.

16a.  With hexane.
16b.  This would include  samples,  check standards,
     duplicates and method blanks.
                                     18a.  This chromatogram will provide the analyst with
                                          the information  for further need of concentration
                                          or clean-up.
                                     18b.  Interferences in the form of distinct peaks and/
                                          or high background will indicate further clean-up
                                          is necessary.

                                     19a.  See next section (Section H).
                                    20a.  If a standard of a concentration at the drinking
                                         water MC has been carried through the procedure
                                         and gives quantitatdble peaks, the analyst can
                                         express his value as less than his detection
                                          limit if he has obtained no peaks in the sample.
                                    20b. Both the sample and duplicate should produce the
                                         same results.
                                    20c. Section J.
                                                                                    /II.G.3.16
                                                                                    [p. 46)
                                                                                       I.B.12
                                                                                       (p- 44)

-------
 kATER  MONITORING  PROCEDURE:   Determination of Chlorinated Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPE RAT ING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
H. Sample Clean-Up
1.  Weigh  out the determined
   amount of florisil.
                           2.  Pour the florisil  into the
                              chroma tographic column.

                           3.  Tap the column lightly to
                              settle the florisil  and  to
                              level  its  surface.

                           4.  Add about  one-half  inch  of
                              sodium sulfate to  the top
                              of the florisil in  the
                              column.

                           5.  Adjust the sample(s)
                              volume to  10 ml.
                           6.  Add 40-50 ml  of petroleum
                              ether to the  top of the
                              column.
                           7. Time the rate of passage
                              of the petroleum ether
                              through  the column.

                           8. Remove the container  used
                              to collect the petroleum
                              ether.
la. Determined from the lauric acid value for this
    batch of florisi1.
lb. Usually about 10 to 20 grams.
Ic. See calculation section (J.2).

2a. The column should  meet the specifications listed
    in the equipment section.

3a. The column should  not be packed so tight as  to
    impede sol vent flow.
                                4a.  This  should  be  the  anhydrous, granular, ACS  grade
                                4b.  Should  be  pre-conditioned  by heating at 400°  C
                                    for  four hours  (See B.ll).
                                5a.  A column  must  also  be  prepared  for  the method
                                    blank,  the  standard check  and each  duplicate
                                    being  run.
                                5b.  Use  hexane.

                                6a.  This is  used to  pre-wet  the  column
                                6b.  The  column  may generate  heat as  it  is wet by the
                                    solvent.  Allow  to  cool  to room  temperature
                                    before  proceeding.
                                6c.  This solvent can be collected in any  type of
                                    container and  need  not be  saved.

                                7a.  This elution rate should be  set  at  about 5 ml
                                    per  minute.
                                7b.  Use  a  10  ml graduated  cylinder  and  a  stopwatch.

                                8a.  This should be done while  a  small level of
                                    solution  is still above  the  sodium  sulfate layer.
                                                                                                             E20-37

-------
 WATER MONITORING PROCEDURE:    Determination of Chlorinated  Hydrocarbon  Pesticides
                                                                                                          E20-38
 OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                    TRAINING
                                                   GUIDE NOTES
H. Sample Clean-Up
   (Continued)
 9. Place a clean 500 ml  K-D
    flask equipped with a 10
    ml  ampul  under the column.

10. Just prior to the exposure
    of the sulfate layer  to
    the air quantitatively
    transfer  the sample
    extract into the column.

11. Rinse container with  10 ml
    of petroleum ether.

12. Add 200 ml  of the 6%  ethyl
    ether in  petroleum ether
    solution.
 9a.
                                                          lOa.
                                                          lOb.
                          13.  Collect the  200  ml
                          H.  Add  200  ml  of  the  15%  ethyl
                              ether-petroleum  ether
                              solution.

                          15.  Immediately  replace  the
                              500  ml K-D  flask with
                              another  clean  one.
Some portion of the petroleum ether will
collected.
be
     The sulfate should not be allowed to dry between
     additions of the sample or future eluates.
     Wash the ampul  with three 5 ml portions of
     petroleum ether and transfer each wash into the
     column.
                                                          11 a.  Add the rinse to the column.
12a. Measure in a graduated cylinder.
12b. If prepared just prior to use by  the directions
     given in the reagent preparation  section,  only
     200 ml will be prepared and can be completely
     transferred to the column.
12c. Add small  portion slowly to bring liquid level
     to the top of the florisil  column then add the
     rest.

13a. Close the  column's stopcock before the sulfate
     layer is exposed to the air.

-------
 WATER MONITORING PROCEDURE;    Determination  of Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/CPFRAT1NG GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
H. Sample Clean-Up
   (Conti nued)
16.  Collect the 15% eluate,
                          17. Just prior to the sulfate
                              layer being exposed add
                              200 ml of ethyl ether.
                          18. Immediately replace the
                              500 ml K-D flask with
                              another clean one,

                          19. Concentrate the elution
                              volumes in their
                              respective flasks.

                          20. Remove the flask from the
                              ampul.

                          21. Rinse lower portion of
                              the flask and glass joint.

                          22. Obtain a  final volume of
                              10 ml.

                          23. Stopper the ampul.

                          24. Inject 5 pi of the
                              fractions.
16a. There will  be a small  overlap;  in this case a
     small portion of the 6% will  be collected in the
     15?= flask.   Again close the stopcock to prevent
     exposure of sulfate layer to the air.

17a. For wastewater or if other pesticides  are to be
     monitored other than those listed in the Interim
     Primary Regulations, a 501 ethyl ether in
     petroleum ether  elution would  be carried out
     at this time followed by the straight  ethyl
     ether elution.
                                19a.  Use  the  procedure  under  Sample  Concentration,
                                     (G.3), Steps  1  through 6.
                                24a.  See F,23a.
                                                                                                              E20-39

-------
 HATER  MONITORING  PROCEDURE:   Determination  of  Chlorinated  Hydrocarbon  Pesticides
                                                                                                        E20-40
 OPERATING PROCEDURES
        STLP  SEQUENCE:
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                TRAINING
                               GUIDE  NOTES
I.  Sample Analysis
1 .  Check instrument
   parameters.
                           2.  Inject 5 yl of standard
                              mixture 2.

                           3.  Determine  if response has
                              changed for the standards.

                           4.  Inject 5 yl of the method
                              blank.

                           5.  Determine  if impurities
                              are present which will
                              interfere.

                           6.  Inject 5 yl of the sample.
                          7. Compare with standard
                             curve (Step I.2.).

                          8. Inject 5 yl of a known
                             standard mixture that is
                             very close to that of the
                             sample.

                          9. Calculate the amounts of
                             the pesticides present.
                                                                                   - 70
                                                                                   - 70
la. Column Temp.
    OV-17 & QF1  - 200°
    OV-210      - 180°
    OV-1         - 180°
    QF1  - SE30  - 200°
Ib. These conditions should be
    analyst during operation.
Ic. If the instrument was off,
    equilibrate  overnignt.
Flow
- 60 ml/min.
     ml/min.
     ml/min,
                                                                                 C - 60 ml/min. -
                                                                                         monitored
Detector
220° C
200° C
200° C
220° C
 by the
                                                                                         turn on and allow to
                                2a.  Reagent  Preparation  Section.
                                3a.  Check  retention  times  as well as peak heights
                                    for  the  known  concentrations.

                                4a.  Continue  the chromatogram until the retention
                                    time for  the last peak has passed.
                               6a. Time the retention time; from first sign of the
                                   solvent until the top of the peak, for each peak.

                               7a. Retention times and numbers of peaks will
                                   indicate further actions.

                               8a. Measure the peak area or peak height in
                                   mil 1imeters.
                               9a. See the Calculation Section.
                                                       VII.1.7.7a
                                                       (p. 47)

-------
 WATER  MONITORING  PROCEDURE;   Determination  of  Chlorinated  Hydrocarbon  Pesticides
 OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING  COALS/SPECIFICATIONS
                                                    TRAINING
                                                   GUIDE  NOTES
J. Calculations
   1.  Laurie Acid Value
   Use the following  steps
   to calculate the  lauric
   acid value.
la.  The calculation formula  is
    Lauric Acid Value = mg lauric  acid
                          gm florisil
                                                                                                     for
                                                      =  200  -  (ml required
                                                               titration x
                                                               mg lauric acid
                                                               ml 0.05N NaOHj
                                    the mg  lauric Ac1d
                                        ml T.05N NaOH"
                                    comes  from  the  standardization of the NaOH
                                    (See Section on Standardization of Sodium
                                     Hydroxide).
   2. Amount of Florisil
      to be Used in the
      Column
1.  Calculate the amount
   of florisil  to be used  in
   the columns.
la.
                                                           Ib.
The calculation formula
Amount of florisil
to be used  =    110
                            is
                                     x 20 grams
                                                   Lauric Acid  Value
                                    The 110  value  is  a  value abritrarily assigned
                                    as  the desired adsorptive capacity.
                                                       Milis,  P.A.,
                                                       Variations  of
                                                       Florisil
                                                       Activity:
                                                       Simple  Method
                                                       for  Measuring
                                                       Adsorbent
                                                       Capacity and
                                                       its  Use 1n
                                                       Standardizing
                                                       Florisil
                                                       Columns;
                                                       JAOAC,  51 29
                                                       (1968)
                                                                                                            E20-41

-------
 WATER MONITORING PROCEDURE:   Determination  of Chlorinated Hydrocarbon Pesticides
                                                                                                         E20-42
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
J. Calculations
   (Continued)
   3.  Calculation of
      Results
1.  Calculate the  micrograms
   of pesticide per  liter  of
   sample.
                                                                               VT * V
la.  The calculation formula  is

    micrograms   _  A x B x V^
       liter"
               where
    A  =  ng standard    (obtained  in  Sample  Analysis
          standard area   Section,  Step  8)

    B  =  Sample Aliquot Area  (obtained  in  Sample
                               Analysis  Section,
                               Step 7)

    V^  =  Volume of total  extract  (i.e.  the  volume
                                    to which  the
                                    extract was
                                    concentrated)
           in microliters
                                   Vi
                                                                  =  Volume of extract injected in microliters
                                                               $  =  Volume of water (sample)  extracted in
                                                                     mil 1 iters.

-------
WATER MONITORING PROCEDURE:  Determination of Chlorinated Hydrocarbon Pesticides

                               TRAINING GUIDE
          SECTION                  TOPIC
            I*                 Introduction
            II                 Educational Concepts - Mathematics
            III                Educational Concepts - Science
            IV                 Educational Concepts - Communications
            V*                 Field and Laboratory Equipment
            VI                 Field and Laboratory Reagents
            VII*               Field and Laboratory Analyses
            VIII               Safety
            IX                 Records and Reports
*Training guide materials are presented here under the headings marked*.
 These headings are used through this series of procedures.
                                                                        E20-43

-------
WATER MONITORING PROCEDURES:   Determination of Chlorinated Hydrocarbon Pesticides
 INTRODUCTION
                                                          Section   I
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
 B.12
When the Interim Primary Drinking Water Regulations
were promulgated they contained the requirement
that all Public Water Supplies be monitored for
pesticide contamination.  Certain nonitoring
frequencies and analytical methods were prescribed.
A level beyond which public notification and other
steps were to take effect was set and termed the
"Maximum Contaminant Level" (MCI).  The MCL's for
the Chlorinated Hydrocarbon Pesticides are as
follows:
                      Endri n
                      Lindane
                      Methoxychlor
                      Toxaphene
                   0.0002 mg/liter
                   0.004 mg/liter
                   0.1  mg/liter
                   0.005 mg/liter.
  E20-44

-------
WATER MONITORING PROCEDURES:   Determination of Chlorinated Hydrocarbon Pesticides
 FIELD AND LABORATORY EQUIPMENT
                                                          Section  V
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
C.I
C.6.b
D.I.3.a
D.3.2
 Use  of high grade  carrier  gases are  recorrmended.
 However, occasionally bad  cylinders  of the Argon-
 Methane gas can be obtained.  Before attaching  to
 the  instrument,a slight  sniff of the gas should be
 taken; if a "fishy" ordor  is noted,  the tank may
 be contaminated.   Use of the gas will produce an
 off  scale peak and very  noisy base line.  If
 contaminated gas is used in the instrument, remove
 from use as soon as it is  determined and replace
 all  traps and purge with a good gas.

 OV-210 may be substituted  for the QF-1 .  The OV-210
 is a purified version of the QF-1 and does not
 bleed as much as the QF-1.

 Because of the many variables inherent in gas
 chromatographs, the column  packing, column oven
 temperature and carrier gas flow rate may have to
 be adjusted to different settings than those given.
 The analyst should strive  to reproduce the retent-
 ion times  given in the body of the paper as guides.
 The two things which must be obtained are
 reproducibility and resolution.   When these are
adequate the system is suitable.

The analyst must know what the detection limit is
 in his  procedure.   The concentrations are so low
 in drinking water that frequently the results
will  be non-detectable.   In this case the analyst
should  express the value as "non-detectable,"
below the  detection limit of
This  detection limit must be at  least below the
Maximum Contaminant Level for the compound as
listed  in  the Interim Primary Drinking Water
Regulations.
                                                                                E20-45

-------
WATER MONITORING PROCEDURES:  Determination of Chlorinated Hydrocarbon Pesticides
FIELD AND LABORATORY ANALYSES
                                                                           Section
                                                                   VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
D.4
F.
F.2
G.Z.lf
G.3.16
Lindane, Endrin, and Methoxychlor are quantitated
using one peak for each.  However, toxaphene has
many peaks and should be quantitated by averaging
the peak height (in millimeters) of as many peaks
as possible.  Use only those peaks which are
identical in the standard and sample by retention
and peak height ratio.

The more peaks used in this average, the closer
this number will come to the true value.  However,
when other chlorinated hydrocarbons are present
the peaks of toxaphene in areas not affected by
the other chlorinated hydrocarbons should be
averaged.  From running the standards separately
these areas can be found.

When the peaks of the other chlorinated hydrocarbons
are influenced by the presence of toxaphene they
should be separated by another technique, i.e.
another absorption column or liquid chromatography
etc.

Different batches of florisil have varying adsorp-
tive capacities.  In order to obtain elution of
the various pesticides in the same fractions, this
adsorptive capacity must be known.  A rapid method
for determining this adsorptive capacity is to
measure the amount of lauric acid adsorbed from
hexane solution by a measured amount of florisil.
This is referred to as the lauric acid value.

This procedure need be carried out only once with
each batch of florisil.   As a new supply is
purchased, the lauric acid value and this procedure
should be determined again, using this same
procedure.

Samples up to 3 liters can be extracted to increase
the sensitivity.  However, larger volumes of
extraction solvent (methylene chloride-hexane)
will be needed.   Use 100 ml portions in place of
the 60 ml.   The separatory funnel size will  also
need to be increased to  a 6 liter size.

Samples containing small  quantities of pesticides
(low nanogram amounts) are concentrated to 1  ml.
Should the concentrations allow, it is possible
to concentrate to 10 ml  for higher values.   In
this case the initial  concentration with the Snyder
Column will  reduce the volume to 5 to 6 ml  and the
  E20-46

-------
WATER MONITORING PROCEDURES:   Determination of Chlorinated Hydrocarbon Pesticides
FIELD AND LABORATORY ANALYSES
                                                          Section  yil
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
G.3.16
(Continued)

I.7.7a
walls of the
give a final
flask and glass joint can be washed to
volume of 10 ml.
Major peaks not matching those of the standards
along with increased peak heights on some peaks
will indicate the presence of interfering
compounds.  If this is the case, the florisil
column clean-up should be used.  If no
interferences are present, identify and quantitate
the peaks.
                                                                               E20-47

-------
                                        TABLE 3





        RETENTION RATIOS OF VARIOUS ORGANOCHLORINE PESTICIDES RELATIVE TO ALDRIN
Liquid Phase


Column Temp.
Argon/Methane
Carrier Flow
Pesticide
Lindane
Aldrin
Endrin
Methoxychlor
Aldrin
(Min absolute)
1.5% OV-17
+
1.95% QF-1
200° C

60 ml/min.
RR
0.69
1.00
2.93
7.6

3.5

5%
OV-210
100° C

70 ml/min.
RR
0.81
1.00
3.56
6.5

2.6

3%
OV-1
180° C

70 ml/min.
RR
0.44
1.00
2.18
5.7

4.0
6% QF-1
+
4% SE-30
200° C

60 ml/min.
RR
0.60
1.00
2.42
4.60

5.6
    All  columns glass,  180 cm x 4 mm ID,  solid support Gas-Chrom 0 (100/120 mesh)
E20-48

-------
                 X
                 O
                               I
                          T
                   25
               20         15         10
                  RETENTION TIME IN MINUTES
ro
c
FIGURE 1. COLUMN PACKING: 1.5% OV-17 + 1.95% QF-1, CARRIER
         GAS: ARGON/METHANE AT 60ML/MIN, COLUMN
         TEMPERATURE: 200 C, DETECTOR: ELECTRON CAPTURE.

-------
                              10          5           0
                       RETENTION TIME IN MINUTES

       FIGURE 2. COLUMN PACKING: 5% OV-210, CARRIER GAS:
                 ARGON/METHANE AT 70ML/MIN, COLUMN
                 TEMPERATURE: 180 C,  DETECTOR: ELECTRON
                 CAPTURE.
E20-50

-------
                    DC
                    O
                    X
                    O
                    i
                   25
                20         15         10
                    RETENTION  TIME IN MINUTES
O
•
en
FIGURE 3. COLUMN  PACKING:  6% QF-1 -I- 4% SE-30, CARRIER GAS:
          ARGON/METHANE AT 60/VU/MIN, COLUMN TEMPERATURE;
          200 C,  DETECTOR: ELECTRON CAPTURE.

-------
IN)
O
I
Ul
PO
                                   O
                                   _J
                                   <
                                                                          z
                                                                          O
                                                        oc.
                                                        O
                                                                        U
                                                                        I
                                                                        CO
                    25
                      FIGURE 4
20         15          10         5
     RETENTION  TIME IN MINUTES

COLUMN PACKING: 3% OV-1, CARRIER GAS:
ARGON/METHANE AT  70 ML/MIN, COLUMN

TEMPERATURE: 180 C, DETECTOR: ELECTRON
CAPTURE.
J
0

-------
                             A PROTOTYPE FOR DEVELOPMENT OF
                             ROUTINE OPERATIONAL PROCEDURES

                                        for the

                 DETERMINATION OF CHLORINATED PHENOXY ACID HERBICIDES
                                      as applied in

                           POTABLE WATER TREATMENT FACILITIES
                  National  Training and Operational  Technology Center
                         Office of Water Program Operations
                        U.S.  Environmental  Protection Agency
CH.PES.lab.WMP.1.11.77
                                                                             E21-1

-------
WATER MONITORING PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides


Operational Procedures

A. Glassware Preparation

B. Reagent Preparation

C. Standard Preparation

D. Instrument Set-Up

   1. Gas Connection
   2. Detector Installation
   3, Carrier Gas Leak Check
   4. Column Conditioning

E. Instrument Calibration

   1. Optimization
   2. Retention Times
   3. Detection Limits

F. Sample Treatment

   1. Pre-Treatment
   2. Hydrolysis
   3. Esterification

G. Sample Analysis

H, Calculations
                                                                         L21-3

-------
     WATER MONITORING  PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides


     Equipment  and  Supply  Requirements

     A.  Capital  Equipment:

        1. Gas  chromatograph  equipped witn

          a. Glass lined  injection  port
          b. Electron capture  detector - tritium or  nickel 63
          c. Recorder - potentiometric strip  chart  (10  in.-25 cm) compatible
             with  the detector

        2. Gas  chromatographic  column  (best  purchased from  gas chromatographic
          supply  house)

          a. Tubing - Pyrex  (180 cm long (6 ft.) x  4 mm ID)
          b. Glass wool - silanized
          c. Solid support - gas chrom Z (100-120 mesh)
          d. Liquid phase -  expressed as weight percent coated  on solid  support

              1) OV-210, 5%
             2) OV-17, 1.5%  plus QF-1, 1.95%

        3. Hot  water bath  - capable  of keeping temperature  at  50°-100°  C

        4. Source of high  quality distilled  water

        5. Rotometers - If the  instrument is not equipped with meters  to  monitor  the
          flows of gases, these should be  purchased as  options.

        6. Analytical  balance - with a 0.1 milligram sensitivity

        7. Trip or platform balance  -  with a 0.1 or  0.01  gram  sensitivity

        8. Oven - capable  of  maintaining  130°  C

        9. Stop watch - capable of  measuring at least 1/2 hour,  the  60  second
           cycle divided to 1/5 second

       10.  Cylinder of Argon-methane (95  +  51) for  use with pulsed mode detector  OR
           Nitrogen - purified  grade,  moisture and  oxygen free,  for  use with  a  DC mode
           detector

       11.  Pressure regulator - two  stage with a CGA 580 fitting for nitrogen  or  a CGA
           350  fitting for Argon-methane

       12.  Filter - for carrier gas, molecular sieve type

       13.  Micro syringes  - 5,  10,  50 yl  sizes

       14.  pH meter (optional)
E21-4

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid  Herbicides


Equipment and Supply Requirements (Continued)

  15. Desiccator

  16. Muffle furnace (optional) - capable of heating to 400°  C

  17. Source of Vacuum

  18. Trap for oxygen

B. Reusable Supplies:

   1, Beaker - 100 ml, I/sample, duplicate, blank,  standard

   2. Cylinders, graduated

      2, 25 ml
      2, 100 ml
      1, 250 ml
      1, 1000 ml

   3. Cylinder, graduated -  glass-stoppered

      1-25 ml/sample, duplicate, blank,  standard

   4. Flasks, Erlenmeyer

      1, 125 ml/sample, duplicate,  blank, standard
      1, 250 ml/sample, duplicate,  blank, standard
      1, 1000 ml/sample, duplicate,  blank, standard
         Erlenmeyer - glass-stoppered
      1, 250 ml/sample = I 19/22

   5. Flasks - volumetric

      4, 100 ml
      6, 10 ml

   6. Funnel, 1 - 50 mm diameter top/sample,  duplicate, blank, standard

   7. Funnel - separatory (with Teflon stopcock)

      1, 2000 ml/sample, duplicate,  blank, standard
      1, 60 ml/sample, duplicate, blank,  standard

   8. Glass stirring rod, about 10  cm long

   9. Glassware brush
                                                                        E2t-;

-------
      WATER MONITORING  PROCEDURE:   Determination  of  Chlorinated  Phenoxy Acid  Herbicides


      Equipment and  Supply  Requirements  (Continued)

        10.  Kuderna-Danish  (K-D),  order  from  Kontes  Glass  Corporation

            5 plus  2/sample,  concentration  ampul,  10 ml  calibrated  5  19/22  female
                   #1(5-70050,  size 1025
                   I/sample,  Snyder col.,  three section,  150  mm  long,  IK503000,  size  121
                   I/sample,  Snyder col.,  one section,  #K569Q01,  size  1-19
                   I/sample,  flask, 250  ml  size,  #K570001
            About 6/pr.,  springs,  2/set-up, #K662750
            I/ampul,  stoppers  for  concentrator ampul, #K850500,  size  19/22

        11.  Pipets
            1 box,  Pasteur, disposable (140 mm long  x 5 mm ID)
            4 - graduated,  1  ml
            5 - volumetric, 1  ml
            3 - volumetric, 2 ml
            2 - volumetric, 10 ml

        12. Pipet bulb

            3 ml size for Pasteur type pipets
            Rubber type for pipets

        13. Reagent bottles (glass-stoppered, storage)
            5 - 150 ml  size
            1 - 500 ml  size
            1 - 1000 ml size

        14. Rack for separatory funnel

        15. 3 - ring stand and clamp

        16. 1 ruler - divides in millimeters

        17. Safety glasses

        18.  1 timer (60 min.)

      C. Consumable Supplies

         1.  1 box aluminum foil

         2.  1 bottle - boiling stones (rinse with hexane)

         3.  1 box detergent

         4.  25  liters distilled water

         5.  1 box glass wool  (filtering grade, acid washed)
E21-6

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides


Equipment and Supply Requirements (Continued)

   6. 1  pack graph paper (arithmetic, 10 x 20)

   7. 1  roll pH paper (for acid pH)

   8. 12 labels

   9. 1  note book (bound)

  10. 1  pencil  or pen

  11. 12 weighing boats, plastic, disposable

  12. 1  bottle  soap solution (any liquid soap mixed with water)

  13. Chart paper for records

  14. Reagents

      a. Acetone - ACS grade
      b. Alcohol  - ethenol,  95%,  ACS grade
      c. Benzene - nanograde, distilled  in glass
      d. Borontrifluoride  -  methanol, esterification reagent,  14% BF3  by  weight*
      e. Ethyl  ether - nanograde, distilled in  glass
      f. Florisil - pesticide residue grade (60-100 mesh),  purchase  activated  at
         1250°  F and store at 130° C
      g. Herbicide standards, reference  grade
      h. Hexane,  nanograde,  distilled in glass
      i. Potassium hydroxide (KOH),  ACS  grade
      j. Potassium iodide  (KI), ACS  grade
      k. Sodium sulfate, ACS, granular
      1. Sulfuric acid,  ACS, concentrated
   *Available already prepared from:   Applied Sicence Laboratories
                                      PO Box 440
                                      State College, PA  16501
                                                                        E21-7

-------
 HATER MONITOR!?^ PROCEDURE;  Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                          E21-8
  OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
A. Glassware Preparation
B.  Reagent Preparation
   1.  Distilled Water
   2.  Sulfuric Acid (25%)
1. Clean all glassware.

2, Wash with soap and water.

3. Rinse with tap water.

4. Rinse with distilled water.

5. Muffle at 400° C for 15 to
   30 minutes.
                           6.  Cool  to room temperature.

                           7.  Store until  used.
                                                          3a.  At least 10 times.

                                                          4a.  At least 10 times,

                                                          5a,  Volumetric glassware should not be muffled.
                                                          5b.  Plastic ware and cap liners for sample containers
                                                              should not be muffled.
                                                          5c.  The glassware may be rinsed with redistilled
                                                              acetone followed by a rinse with pesticide quality
                                                              hexane in place of the muffling.
                               7a. Store inverted or cover mouth with aluminum foil.
                               7b. Sample containers should be stored capped.
1. Distill water.
1. Add 50 ml of water to a
   100 ml graduated cylinder.

2. Add 25 ml of concentrated
   sulfuric acid.

3. Cool to room temperature.

4. Add water to the 100 ml
   mark.
la.  Use an all  glass still.
Ib.  Extract a volume of distilled water,  equal  to the
    sample size used, to check purity.   This reagent
    blank should be analyzed with each  set of samples
                                                           2a.  Measure in a  25 ml  graduated cylinder.
                                                           2b-  CAUTION:   Temperature will  rise.

-------
 HATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   3.  Sodium Sulfate
      Solution (5%)
      (Na2S04)
   4.  Potassium Iodide
      (10%)  (KI)
5.  Transfer to a reagent
   bottle.

6.  Store in freezer section
   of a refrigerator.

1.  Weigh out 5.0 grains of
   sodium sulfate.

2.  Transfer to a glass-
   stoppered reagent bottle.

3.  Measure  out 100  ml  water.

4.  Use a small amount  of water
   to wash  the weighing boat.

5.  Add remaining water to a
   reagent  bottle.

6.  Mix and  label.

1.  Weigh out 10.0 grams of
   potassium iodide.

2.  Transfer to a glass-
   stoppered reagent bottle.

3.  Measure  out 100  ml  water.

4.  Use small amount of water
   to wash  weighing boat.

5.  Add remianing water to
   reagent  bottle.

6.  Mix and  label.
la. Use a trip balance and a plastic weighing boat.
Ib. This is not the acidified sodium sulfate (B.8).

2a. A 150 ml size.
                                                           3a.  In a graduated cylinder.

                                                           4a.  Add washing to reagent bottles.
la. On a trip balance in a plastic weighing boat.


2a. A 150 ml size.


3a. In a graduated cylinder.

4a. Add wash to reagent bottle.
                                                                                                           E21-9

-------
 WATER MONITORING PROCfDLRE:   Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                           E21- 10
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B.  Reagent Preparation
   (Continued)
   5.  Potassium Hydroxide
      (w:v) (KOH)
   6.  Diethyl  Ether
1. Weigh out 37.0 grams of
   potassium hydroxide
   pellets.

2. Transfer to a glass-
   stoppered reagent bottle.

3. Measure out 100 ml water.

4. Use small amount of water
   to wash the weighing boat.

5. Add remaining water to
   reagent bottle.

6. Mix and label.

1. Test for peroxides.
                           2.  Rinse a  25 ml  glass-
                              stoppered graduated cylin-
                              der with ethyl ether.

                           3.  Add 10 ml ether  to the
                              cylinder.

                           4.  Add 1 ml freshly prepared
                              potassium iodide (KI)
                              solution.

                           5.  Mix by inverting.
                                                           2a.  A  150  ml  size.


                                                           3a.  In  a graduated  cylinder.

                                                           4a.  Add washing  to  reagent  bottle.


                                                           5a.  CAUTION:   Temperature will  rise.
la. Purchase distilled in glass  or nanograde.
Ib. Ether must contain 2% alcohol and be free of
    peroxides as follows.

2a. Discard the rinse.
                                4a.  Use a 1.0 ml  volumetric pipet.
                                5a.  Two or three times.   Open stopper to allow
                                    pressure out.

-------
 WATER M.pNJTORINjj PROCEDURE:    Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PRCCLDURiS
        STEP SEQUENCE
INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                          TRAINING
                                                         GUIDE NOTES
B. Reagent Preparation
   (Continued)
   7,  Ethyl Ether -
      Hexane Mix (1:1
   8.  Sodium Sulfate
      (Acidified)
6, Let stand one minute.

7, No yellow color should be
   observed in either layer.
                                                           7a.
                                                           7b.
     If yellowing occurs, the peroxides can be decomposed
     by adding 40 ml of 30% ferrous sulfate solution
     to each liter of ether.  (CAUTION:  Reaction may
     be violent if ether contains a high concentration
     of peroxides).  Then distill.
     If this is needed, an alternate source for a
     better product should be sought.
                           8.  Discard tested ether.

                           9.  Add 20 ml  of 95% ethanol
                              to one liter of ether.

                           0.  Store in refrigerator.
1. Mix equal amounts of the
   hexane and ethyl  ether,

2. Store in glass-stoppered
   reagent bottles.

1. Weigh out 100 grams of
   sodium sulfate.

2, Add ether to just cover  the
   sodium sulfate.

3. Add 0.1  ml of concentrated
   sulfuric acid.
                           4.  Swirl  to mix.
lOa.

lOb.


 la.


 2a.
 la.
 Ib.
                                                           3a.

                                                           3b.
Storage of all flammable solvents should be in
an explosion proof refrigerator.
Store in reagent bottle glass-stoppered (1 liter
size).

Example:  Add 250 ml  hexane to 250 ml ether.
                                                               Use a 500 ml  size.
Use a trip balance.
Use a 250 ml  Erlenmeyer flask.
     Use caution when working with concentrated
     acids.
     Use safety glasses.
                                                                                                            E21- 11

-------
 WATER MONITORING PROCEDURE:   Determination of Chlorinated  Phenoxy  Acid  Herbicides
                                                                                                            E21-12
 CPtRATING PROCEDURES
        STLP SEQUENCE
     INFRMAT ION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
B,  Reagent Preparation
   (Continued)
 5. Remove ether with vacuum.

 6. Weigh 1.0 gram  of  the
   dried sodium sulfate.

 7. Add 5 ml of distilled
   water.

 8. Swirl to mix.

 9. Check pH of the solution.
                          10.  Discard the solution.

                          11.  Store the remaining sodium
                              sulfate at 130°  C  in glass
                              bottles in an oven.
   Standard Preparation
   1.  Stock 2,4  Dichloro-
      phenoxyacetic  acid
      (2,4-D)
1.  Weigh out 0.100 grams  of
   2,4-D.
                          2. Measure out 60 ml of
                             ethyl ether.

                          3. Pour 40 ml of ethyl ether
                             into a glass-stoppered
                             100 ml volumetric flask.
                                                           6a, On a trip balance,
                                                           6b. In a 50 ml beaker.

                                                           7a. Use a 5 ml graduated pipet.
 9a. Solution should have a pH below 4.
 9b. If not, repeat steps 2 through 9.
 9c. Use a pH meter or paper that can distinguish
     a pH below 4.

lOa. Prepared in step 9.

lla. Or activate overnight at 130° C before use,
     storing in a desiccator.
 la.  Prepare standard as shown in the accompanying
     chart,
 Ib.  Weigh on an analytical  balance,
 lc.  Use a plastic weighing  boat.
 Id.  Carry out this procedure in a hood or well
     ventilated area.

 2a.  In a 100 ml  graduated cylinder,
 2b-  CATION:   Ether is  extremely flammable.

-------
 WATER MONITORING  PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
        STEP  SEQUENCE
    INFORMATION/OPERATING COALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C.  Standard Preparation
   (Continued)
   2.  Stock Silvex
   3. Working Standard
      2,4-D
   4.  Working Standard
      Silvex
4. Transfer the 2,4-D into
   the volumetric flask.

5. Use the remaining ether
   to wash the weighing
   boat.

6. Dissolve the 2,4-D.

7. Dilute to the mark with
   hexane.

8. Stopper and label.

1. Perform steps 1  through 8
   above  using si 1 vex.

1. Add about 50 ml  of the
   ethyl  ether:hexane mixture
   (reagent 7)  to a  glass-
   stoppered 100 ml  volumetric
   flask.

2. Pipet  10.0 ml of the stock
   2,4-D  into the flask.

3. Dilute to the mark with
   the ether:hexane mixture.

4. Stopper and label.

1. Add about 50 ml  of the
   ethyl  ether:hexane mixture
   (reagent) to a glass-
   stoppered 100 ml  volumetric
   flask.
                                                           7a.  Solution contains 1  mg/ml
la.  Solution contains 1  mg/ml
la. Just estimate this amount.
                                                           2a.  Use a 10.0 ml  volumetric pipet.
                                                           3a.  Solution contains 100
                                                                                     ml
                                                                                                            E21-13

-------
 HATER  MONITORING  PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                          E21-14
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OFtRATING GOALS/SPECIFICATIONS
  TRAINING
 GUIDE  NOTES
C.  Standard Preparation
   (Continued)
   5.  Esterification of
      Standard for
      Chromatography
      2,4-D
2, Pipet 1.0 ml of the stock
   si 1 vex into the flask.

3. Dilute to the mark with
   ether:hexane mixture.

4. Stopper and label.

1. Turn on water bath and
   steam bath.

2. Pipet 1.0 ml of the working
   standard into a 10 ml
   concentrator ampul.

3, Add 0.5 ml of Benzene.

4. Insert a two chamber
   evaporative column.

5. Place assembly into ring
   stand and support with
   clamp.

6. Place over steam bath.
                          7. Lower volume to 0,4 ml.

                          8. Remove from steam bath and
                             allow to cool.

                          9. Remove the evaporative
                             column.
                                                           3a.  Solution  contains  10
                                                                                    ml
la. Allow water bath to reach 50° C,
Ib. Allow steam bath to boil.

2a. Use a 1.0 ml volumetric pipet.
2b. See equipment  section,  K-D concentrator ampul


3a. Use a 1 ml graduated pipet.

4a. See equipment section,
4b. Be sure to attach springs.
VII.5.1
(p. 38)
                                                           6a. The  height  above  the  steam bath will  have  to  be
                                                              adjusted  to speed  up  or  slow down  the evaporation.
                                                              The  analyst should  not allow the material  to  go
                                                              to dryness.

-------
 WATER HnNITORING  PROCEDURE:   Determination of Chlorinated Phenoxy Acid  Herbicides
 OPERATING PROCEDURES
         STEP  SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
C. Standard Preparation
   (Continued)
   6. Esterification
      of Standard for
      Si 1 vex
      Chroma tography

   7. Esterification of
      Standard Mixture 1
10.  Add 0.5 ml  of the
    borontri f1uori de-methanol
    reagent.

11.  Replace the evaporative
    column.

12.  Lower into  a preheated
    50° C water bath.
13.  Heat for 30 minutes.

14.  Remove from water bath.

15.  Allow to cool  to room
    temperature.

16.  Remove column.

17.  Wash walls of  concentrator
    ampul  with benzene  until  a
    total  volume of 5 ml  is
    reached.

18.  Stopper and mix.

 1.  Repeat steps 1-7 above
    using silvex working
    standard in place of  the
    2,4-D. in step 2.

 1,  Pipet 2.0 ml of each  work-
    ing standard (for 2>4~D.
    and silvex) into a  10.0  ml
    concentrator ampul.
                                                          12a,  This  is  not a  steam bath.
                                                          12b.  The water bath should  be  preheated  to  50°  C  and
                                                               checked  with a thermometer.
                                                          17a.  Solution  contains  200 ng  esterified  2  4-D  per
                                                               10  pi.                                '
la.  These esterified standards are at the level  of
    the sample containing the MCL's after it has
    been concentrated.   That is the MCL in 5 ml.
Ib.  Solution contains 40 ng silvex/ 10 pi.

la.  Use two 2 ml  volumetric pipets.
                                                                                                           E21-15

-------
 WATERMONITORING  PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                         E21- 16
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFORMATION/OPERATING GOALS/SPECIFICATIONS
                                                     TRAINING
                                                   GUIDE  NOTES
C.  Standard Preparation
   (Continued)
2, Add 1.0 ml of benzene.

3. Insert a two chamber
   evaporative column.

4. Place assembly into ring
   stand and support with
   clamp.

5. Place over steam bath.
                           6.  Lower  the  volume  to  1.0 ml.

                           7.  Remove from  steam bath.

                           8.  Remove evaporative column.

                           9.  Add  1.0 ml of  the boron-
                              trifluoride  rnethanol
                              reagent.

                          10.  Replace the  evaporative
                              column.

                          11.  Lower  into a preheated
                              50°  C  water  bath.

                          12.  Heat for 30 minutes.


                          13.  Remove from water bath.

                          14.  Allow  to cool.
2a.

3a.
Use a 1 ml graduated pipet.

Attach springs.
                                                           5a.
    The height above the steam bath will  have to be
    adjusted to speed up or slow down the evaporation
    rate.   The analyst should not allow the solution
    to go  to dryness.
                               12a.
    The bath  must be  at temperature before lowering
    the assembly.

-------
 HATER MONITORING PROCEDURE:  Determination of Chlorinated Phenoxy Acid  Herbicides
 OPERATING PROCEDURES
         STEP  SEQUENCE
JHFORMATION/CFERAriNG GOALS/SrECIFICATIONS
 TRAINING
GUIDE NOTES
C. Standard Preparation
   (Continued)
   8. Dilutions
15.  Remove evaporative column.

16.  Wash walls of concentrator
    ampul  until  a total  volumn
    of 5.0 ml  is reached.

17.  Stopper and mix.

 1,  Prepare dilutions  of this
    mixture labeled Mix 2  and
    3 as shown in Table 1.
                                                          16a.  Solution contains  400 ng  2,4-D/lO  yl  and
                                                               40 ng silvex/10 pi.
                                                                                                           E21-17

-------
         WATER MONITORING PROCEDURE;  Determination of Chlorinated Phenoxy Acid Herbicides


                                     STANDARD PREPARATION

                                           TABLE I
              2,4-D

               100 mg
         Stock 100 ml
       Working   ml
1  ml
               10/100 dilution
              0.1 mg     100 yg     2 ml 2,4-D   -
                     or
   ml
                1/5 dilution
Esterification  100 yg     200 ng
Jingle  Std.  1    5 ml   °   10 pi
                1/2 dilution
               100 ng
   Standard  2    10 yl
                1/2 dilution
               50  ng
   Standard  3   10  pi
                               2.4,5-TP - Silvex

                               1 mg _ 100 mg
                               1 ml ~ 100 ml   Stock
                                                   1/100  dilution
             10
                 2 ml  Silvex  ]p ]
                               ml
                      2/5  dilution
                        and
                     Esterification
                     of Mixtyre
400 ng    M.  ,   40 nc
-iQ-J    Mlx ]   Tori
                                             1/2 dilution
                                                       _
                                                     10
                                             1/2 dilution
                                   100 ng
                                    TO yl
_ 0.01 mg_
    ml    Working
                                                   1/5  dilution
                              20 ng    10 yg   Esterification
                              TOyl y   5 ml   Single Std.  1
                                                   1/2  dilution
                                                   20  ng
                                                   10  yl    Standard 2
                                                   1/2 dilution
                                                   50  09
                                                   T0~yl   Standard 3
                      F11x  3
                  10 ng
                  10 yl
    E21-18

-------
 WATER  MONITORING  PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
        STEP  SEQUENCE.
    Itif CRMATlQN/OP'iRAT I KG  GOAL S/SPEC IFICAT IONS
  TRAINING
 GUIDE  NOTES
D. Instrument Sep-Up
   1.  Gas Connection
   2. Detector
      Installation
   3. Carrier Gas Leak
      Check
1.  Remove cylinder cap from
   cylinder of carrier gas.
                           2.  Install  the pressure
                              regulator.
                           3. Connect the cylinder to
                              the instrument.
1.  Install  the electron
   capture  detector.
1.  Open master valve on
   instrument.

2.  Adjust regulator control
                           3. Close carrier gas control
                              valve on the instrument.

                           4. Turn off master cylinder
                              valve.

                           5. Observe gage on the
                              cylinder.
la.  Tank should be chained to wall  or lab bench.
Ib.  Use Argon-methane for pulsed mode detector or
    use nitrogen for a detector operated in a DC mode.
    Consult the instrument manufacturer's manual.

2a,  The regulator should have a CGA 350 fitting for
    Argon-methane or a 580 fitting  for nitrogen.
2b.  Fitting should be tight to prevent leaks.

3a.  Use Teflon tape on all metal threads to prevent
    leaks.
3b.  Polyethylene (1/8" diameter) tubing can be used.
3c.  If plastic tubing is used, nylon ferrules should
    be used with connector fittings.

la.  This can be of tritium or nickel 63 type.
Ib.  Requires license from Atomic Energy Commission.
Ic.  See the manufacturer's manual on procedures for
    installation.
                                                           2a.  To about 65 psig and allow to stabilize
                                                               (about 1  minute).
                                5a.  Pressure should not drop more than a few psig.
                                5b.  If pressure does drop,  use soap solution to
                                    locate leak.
                                5c.  Correct leak  and check  again.
V.D.I
(P. 37)
                                                                                                            E21-19

-------
 UATER MONITORING PROCEDURE:  Determination  of Chlorinated  Phenoxy  Acid  Herbicides
                                                                                                          E21-20
 OPERATING PROCEDURES
        STEP SEQUENCE
      INFORMATION/OP[RATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
D.  Instrument Set-Up
   (Continued)
   4.  Column Conditioning
6.  Use soap solution and check
   for leaks at the injector
   port and column connections,

1.  Install  packed column in
   oven by  connecting  only the
   column inlet.
                           2.  Start  a  flow of  carrier  gas
                              through  the detector only.
                          3. Begin a low flow of
                             carrier gas through the
                             column.
                          4. Wait 5 minutes.

                          5. Turn on the power to the
                             oven.

                          6. Adjust column temperature tc
                             near maximum recommended
                             temperature for the liquid
                             phase being used.

                          7. Continue heating for 2
                             hours.
 la. These columns may be purchased and meet the speci-
    fications as listed under  the equipment section.
 lb. Column conditioning is essential to eliminate
    column bleed of packing materials and to provide
    acceptable analysis.
 Ic. Do not connect the column  to the detector.
 Id. If in doubt as to column installation, refer to
    the manufacturer's manual.

 2a. Use the purge gas line or  in a dual column oven
    by connecting an unpacked column to the detector.
 2b. In some systems it may be necessary to temporarily
    connect the carrier gas to the air or hydrogen
    inlet in order to get a flow of carrier gas to
    the detector.   The manufacturer's manual  should
    be consulted.

3a. Less  than 60 ml/min.  (^0-50).
3b. To remove oxygen and other trapped gasses.
3c. This  will  be two separate flows.   The column should
    not be connected to  the detector.
                              5a. Consult manufacturer's manual for location
                                  and necessary steps.

                              6a. Still no flow of carrier gas through the column.
                              5b. Column                           Max. Temp. °C
                                  OV-2'TO (5%)                           -275	
                                  OV-17 (1.5%) and QF-1  (1.95%)          250
                                                         V.D.6.6b
                                                         (p. 37)

-------
 WATER  MONITORING  PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PRCCFDURES
D. Instrument Set-Up
   (Continued)
    STEP SEQUENT
                            9.
Reduce temperature to
about 40° C below maximum
temperature.

Allow column temperature
to equilibrate.
                           10.  Check carrier gas  flow.
                           11.  Allow to remain at temper-
                               ature and flow for one
                               hour.

                           12.  Increase temperature to
                               about 20°C above operating
                               temperature.
                           13.  Continue the same flow of
                               carrier gas.

                           14.  Allow to equilibrate for
                               24 to 48 hours.

                           15.  Turn off oven and allow to
                               cool  to room temperature.
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
fla. See table above (6b).
9a. Minimum of 30 minutes.
9b- CAUTION;  Bleed off of the liquid phase will
    occur if the column temperature is not fully
    equilibrated.
                           lOa.
                           TOb.
                                                          lOc.

                                                          lOd.
                           12a.
                           12b.
    At about 50 ml/minute.
    If the instrument has not come equipped with
    rotometers to monitor the flow rate of the
    carrier gas, this should be purchased as an
    option.
    The pressure regulator  on the cylinder should be
    set at 65 psig.
    The electron capture must be installed but the
    column connection should not be made.
    This is operating temperature not maximum
    temperature.
    The temperatures would be:
    220° C for OV-17 and QF-1
    200° C for OV-210
                                CAUTION:   Do not exceed maximum recommended
                                temperature.  See 6b this  section.
 TRAINING
GUIDE NOTES
                                                                                                           £21-21

-------
 HATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                          L21- 22
 OPERATING PROCEDURES
D. Instrument Set-Up
   (Continued)
E.  Instrument Calibration
   1.  Optimization
         STEP SEQUENCE
16. Connect column to the
    detector.

17. Adjust the carrier gas
    flow to the method flow
    rate.

18. Turn on oven and adjust
    to method column
    temperature.

19. Allow the instrument to
    equilibrate at least one
    hour.
 1.  Check instrument operating
    conditions  after the in-
    strument has  been set up.
 2.  Inject 5
    mixture.
                                       yl of  standard
                          3. Adjust the operating
                             parameters to achieve
                             optimum results.
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
16a, Check connection with soapy solution.
                                                          17a.  70 ml  per minute.
                                                          18a.  185°  C.
                                                          19a.  Preferably overnight.
la.

Ib.
Ic.
     The oven temperature should  be stabilized  at  the
     method  temperature  of 185° C.
     The flow rate  constant on 70 ml/minute.
     The column  conditioned.
 Id.  The  system checked  for  leaks.

 2a.  This standard  mixture was  prepared  from  the
     stock and  esterffied.
 2b,  The  actual  volume of material  injected should  be
     kept constant.   That is, the  same volume for
     standards  and  samples.
 2c.  The  standard mixture 2  can  be  used  to optimize
     the  instrument  initially and  thereafter  to
     monitor  its performance.

 3a.  Best resolution  (separation of peaks) and
     retention  times  can be  achieved by  adjusting the
     column temperature and/or the carrier gas flow
     rate.
                                                         TRAINING
                                                       GUIDE  NOTES

-------
WATER MONITORING PROCFDURF:   Determination of Chlorinated  Phenoxy Acid  Herbicides
 OPERATING PROCEDURES
        5TFP  SEQUENCE
    INFORMATION/OPIATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
E.  Instrument Calibration
   (Continued)
   2.  Retention Time
      Determination
   3.  Detection Limit of
      the Instrument
1.  Inject  5  ul  of standard  1
   of 2,4-0 and  Silvex one at
   a time.
                              With a stop watch,  measure
                              the time elapsed  between
                              the first appearance  of the
                              solvent peak and  the  peak
                              of the standard.
                           3.  Repeat steps
                              more times.
                1  and 2 six
1.  Inject  5  ul  of the
   herbicide  standards.
                           2.  Continue with  standards  2
                              and 3 of the  2,4-D  and
                              then si 1 vex and  determine
                              the detection  limit of
                              each herbicide.
3b. Compare results with the standard chroma tograms
    attached (Figure I).
3c. Caution should be taken to allow the instrument
    to equilibrate after any changes,
3d, Optimum results would include good separation of
    peaks, good sensitivity, and good reproducibility.

la. The individual standards of the 2,4-0 and silvex
    should be run.  Not mixtures.
Ib. Injection is a technique which must be learned
    and practiced in order to make accurate and
    reproducible injections.

2a. This time is called the retention time.
2b. After the retention times of the single standards
    have been obtained,  the mixtures should be
    chromatographed to determine that no changes in
    retention times occur.

3a, At least 7 repeat times should be obtained and
    the mean value obtained for each peak.
3b. If in subsequent injections the retention times
    vary signficiantly,  the system parameters
    (E.l.la-d) should be checked over.

la. Begin with standard 1 Table 1 of each herbicide
    and continue with standard 2 and 3.
Ib. The standards must be esterified.

2a. Always inject  5  ul,
2b. The detection limit is that concentration of a
    standard whose peak is twice the highest peak
    caused by instrument noise and is run at the
    most sensitive setting of the instrument.
                                                                                                                  VII.E.Z.lb
                                                                                                                  (p. 38)
                                                                                       ¥.E.3.2b
                                                                                       (p. 37)
                                                                                                             E21-23

-------
 WATER MONITORING  PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                         E21-24
 OPERATING PRCCEPURES
                                    INFORMATION/OPERATING GOALS/SPtCIFICATIONS
                                                         TRAINING
                                                        GUIDE  NOTES
E. Instrument Calibration
   (Continued)
F.  Sample Treatment
   1.  Pretreatment
   2,  Extraction
   The peak height measured
   in millimeters should be
   plotted against the known
   concentrations to produce
   a standard curve.
1.  Blend the sample.
                           2.  Adjust pH  to  2  or  below.
1.  Measure out 1  liter of
   sample.
                           2. Transfer to a 2  liter
                             separatory funnel.

                           3. Add 150 ml of ether to the
                             separatory funnel.
3a. The curve produced can be used to select a con-
    centration that produces a response (peak height)
    close to a unknown sample's peak height.  This
    concentration will be used to calculate the
    unknown concentration.  See the calculation
    section.
la. If suspended matter is present.  This is usually
    not required for drinking waters.

2a. Use concentrated sulfuric acid.
2b. Use a pH meter or indicator paper to measure pH.
2c. Usually not necessary for drinking waters.
la. Experience with the sample source will indicate
    if smaller volumes of sample can be used.  If
    smaller volumes are used, they should be diluted
    to one liter before extraction.
Ib. A volume of distilled water equal to the sample
    volume used should be carried through the pro-
    cedure every time to act as a method blank.
Ic. Analyze one duplicate sample with each run  as a
    quality control check.
Id. One ml each of working dilution of 2,4-D and
    si 1 vex should be diluted to 1 liter and should be
    carried through the procedure to check on the
    standard curve and to assure the analyst of
    proper operation of the wet and instrumental
    sections of the method.

2a. Use a 2 liter funnel in order to have room  to
    shake.

3a. If the sample container has been emptied, use the
    ether to wash the container and cylinder used for
    transfer.

-------
HATER  MONITORING  PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
STEP SLQULNU
INFORMATION/OPERATING  GOALS/SPECIFICATION'S
 TRAINING
GUIDE NOTES
F. Sample Treatment
   (Continued)
                           4.  Shake virorously for one
                              minute.
                           5,  Place in holder and allow
                              phases to separate for at
                              least 10 minutes.
                           6.  After separation  of the
                              phases is completed,  drain
                              the water phase into  a
                              one-liter Erlenmeyer  flask
                           7.  Add 2 ml  of the 37% aqueous
                              potassium hydroxide solu-
                              tion (reagent 5)  to a
                              250 ml  ground glass
                              stoppered Erlenmeyer flask.

                           8.  Drain the ether layer from
                              the separatory funnel  into
                              the Erlenmeyer containing
                              the potassium hydroxide
                              solution.

                           9.  Close the stopcock and pour
                              the aqueous phase into the
                              separatory funnel.
                        3b. Measure in a 250 ml graduated cylinder.

                        4a. Pressure may build up in the separatory funnel.
                            Invert, with stopper tightly in place, and open
                            stopcock slowly to relieve pressure.  Then con-
                            tinue with shaking until one minute is up.  Re-
                            lieve pressure several  times during shaking.
                        5a. Holder may be a ring and stand or some form of
                            separatory funnel  rack.   Places should be pro-
                            vided for the blank, duplicate standard and all
                            samples.

                        6a. If emulsions form and prevent adequate separation,
                            drain the aqueous  layer that has separated.  In-
                            vert the separatory funnel, and shake rapidly.
                            Vent the funnel frequently to prevent excessive
                            pressure buildup.
                        6b. Assure no water phase remains with the ether.
                            The water is highly acid and step 7 is to assure
                            an alkaline condition in the ether.
                        6c. Remove the stopper before opening the stopcock.

                        7a. The 5 size of the  flask should be capable of
                            fitting the three  ball  Snyder columns.
                                                                                                            E21-25

-------
 MTER  MONITORING  PROCEDURE:  Determination of Chlorinated Phenoxy Acid  Herbicides
                                                                                                         E21-26
 OFERATING PROCEDURES
         STEP SEQUENCE
INFORMATION/OPERATING  GOALS/SPECIFICATIONS
                                                         TRAINING
                                                        GUIDE NOTES
F.  Sample Treatment
   (Continued)
   3. Hydrolysis
10. Add 50 ml of ether.
                          11.  Stopper and  shake  vigor-
                              ously  for one  minute.
12. Place in holder and allow
    phases to separate for at
    least 10 minutes.

13. After separation of the
    phases is complete, drain
    the water phase in the
    one liter Erlenmeyer flask
    used in step 6.

14. Combine the other phase
    with the first ether
    extract.

15. Repeat steps 9-14 with a
    third extraction of 50 ml
    of ether.

 1. Add 15 ml of distilled
    water and one small boilinc
    stone to the flask con-
    taining the ether extract.

 2. Insert a 3 section Snyder
    column into the flask.

 3. Suspend flask and column
    over a steam bath.
lOa.  Use the ether to rinse out the flask used to
     contain the aqueous layer.

lla.  Occasionally invert the separatory funnel and
     relieve the pressure by slowly opening the
     stopcock.
                                                          13a. Assure no water  phase  remains  to  be  collected
                                                              with  the ether.
 la.  Use  a  25  ml  graduated  cylinder.
                                                          3a. Support with ring stand and clamp.

-------
MATER MONITORING PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
       STEF  SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Sample Treatment
   (Continued)
.  Heat until  ether has
  evaporated.
                           5,  Continue heating for a
                              total  of 60 minutes.

                           6.  Remove and allow to cool.

                           7.  Transfer the water con-
                              centrate to a 60 ml
                              separatory funnel.

                           8.  Add 20 ml  of ether to the
                              separatory funnel,
                           9.  Stopper and shake one
                              minute.

                          10.  Place in holder and allow
                              phases  to separate for at
                              least 10 minutes.

                          11.  Drain aqueous  layer into  a
                              100 ml  beaker.

                          12.  Drain off and  discard the
                              ether.

                          13.  Return  the water to the
                              separatory funnel.

                          14.  Add 20  ml  ether and repeat
                              steps 9-13.
4a. This should be carried out in a hood,
4b. Do not allow flame in or near the hood.
4c. Ether fumes are extremely flammable.
                               7a.  Use  the  same  support as  with  other separatory
                                   funnels.
                               8a.  Use  a  25 ml  graduated  cylinder.
                               8b.  Because  the  solution  is  basic,  the  herbicides
                                   will  remain  in  the aqueous  phase.

                               9a.  Vent  pressure  in  separatory funnel.
                              14a.  Again  discard  the  ether.
                                                                                                            E21-27

-------
 WATER  MONITORING  PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                         E21-28
 OPERATING PROCEDURES
         STEP SEQUENCE
     INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Sample Treatment
   (Continued)
 15. Rinse  the  TOO ml beaker
    with a  small amount of
    water  and  add to the
    separatory funnel,

 16. Add 2  ml of cold (4° C)
    25% sulfuric acid  (reagent
    2) to  the  separatory
    funnel.

 17. Add 20  ml  of ether to the
    separatory funnel.

 18. Shake  one minute.

 19. Allow  to stand and sepa-
    rate for at least 10
    minutes.

 20. Drain  the aqueous layer
    into a  TOO ml beaker.

 21. Add 0.5 grams of acidified
    anhydrous sodium sulfate
    (reagent 8) to a 125 ml
    Erlenmeyer flask.

 22. Collect the ether in the
    125 ml  Erlenmeyer flask.

 23. Return the water layer to
    the separatory funnel.

24. Add 10 ml ether.

25. Repeat steps  17-22.
15a.  Use about 2 ml.
                                                          16a.  This  reagent  should  have  been  stored  in  the
                                                               freezer  section  of a refrigerator  (B.2.6).
                                                          16b,  This  will  acidify  the solution;  the herbicides
                                                               are now  soluable in  the organic  phase.
                                                         22a. The herbicides are  in the ether.
                                                         23a. Close the stopcock.

-------
WATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
F.  Sample Treatment
   (Continued)
   4.  Esterification
         STEP  SEQUENCE
26.  Extract with another 10 ml
    volume of ether.
27.  Collect all  ether ex-
    tractions in the same
    125 Erlenmeyer (step 22).

28.  Allow the ether extract to
    remain in contact with the
    sodium sulfate (step 21)
    for 2 hours.

29.  Turn on water bath and
    heat to 50°  C.

 1.  Connect a 10 ml graduated
    ampul to a 250 ml Kuderna-
    Danish (K-D) evaporative
    flask.

 2.  Clamp flask  and ampul to a
    ring stand.

 3.  Plug the stem of a small
    funnel with  glass wool.

 4.  Position the funnel  stem
    into the K-D flask.

 5.  Transfer the ether extract
    from the Erlenmeyer flask
    through the  funnel and
    into the K-D flask.
     INFORMATION/OPERATING GOA!S/SPECIFICATICNS
26a. This will  be three extractions, one with 20 ml
     and two with 10 ml of ether.
26b. On final extraction, rinse the 100 ml beaker used
     to collect the aqueous phase with the ether and
     then add the ether to separatory funnel.
 la. Attach springs.
                                                           3a. Should have been washed with acid.
                                                           5a. Use a glass stirring rod and crush any caked
                                                               sodium sulfate.
                                                           5b. The sodium sulfate may be transferred to the
                                                               funnel.
 TRAINING
GUIDE NOTES
                                                                                                            E21-29

-------
 WATER  MONITORING  PROCEDURE:   Determination of Chlorinated  Phenoxy Acid Herbicides
                                                                                                         E21-30
 OPERATING PROCEDURES
         STEP SEQUENCE
INFORMATION/OPERATING GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F.  Sample Treatment
   (Continued)
 6. Wash the Erlenmeyer flask
    and sodium sulfate with
    liberal amounts of ether.

 7. Add 0.5 ml benzene.

 8. Position flask above steam
    bath and evaporate to about
    5 ml.

 9. Rinse the flask with 2 ml
    of ether.

10. Remove ampul from flask.

11. Insert the two section
    Snyder microcolumn into
    the ampul.

12. Return to steam bath and
    concentrate to about
    0.4 ml.

13. Remove from steam bath and
    allow to cool.

14. Remove column.

15. Add 0.5  ml  of boron tri-
    fluoridemethanol  reagent.

16. Return two chamber column
    to  the ampul.

17. Lower into a preheated
    50° C water bath.
                                                           7a.  Use  a  1  ml  graduated  pi pet.

                                                           8a.  A  condenser column  is  not  necessary.
                                                          lla. Attach  springs.
                                                          12a. All ether should have been evaporated.
                                                         17a. Support by a ring stand and clamp.
                                                         17b. This is not the steam bath.

-------
 lyAT£_R_HQ_N I TORI NG PROCE PURE ;   Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
F,  Sample Treatment
   (Continued)
         STFP SEQUENCE
18. Heat for 30 minutes.

19. Remove from bath and cool
    to room temperature.

20. Remove column.

21. Add enough sodium sulfate
    solution (reagent 3} to
    position the interface,
    between the benzene and
    aqueous sodium sulfate,
    into the neck of the K-D
    ampul,

22. Stopper the ampul and
    shake vigorously for about
    one minute.

23. Allow to stand and
    separate.

24. Plug a disposable Pasteur
    pipet with glass wool,

25. Add florisil absorbant to
    the column.

?6. Place on top of the flori-
    sil some granular sodium
    sulfate.

27. Support the mini-column.

28. Position mini-column into
    the neck of a 10 ml  gradu-
    ated K-D ampul.
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
18a.  The column acts as an air cooled condenser.
                                                          21a. Usually about 4 to 5 ml.
                                                          23a.  About three minutes.
                                                          25a.  Add enough to provide about 2.0 cm height (in the
                                                               large diameter section on the pipet).

                                                          26a.  Enough to provide a sodium sulfate length of
                                                               2.0 cm.
                                                          27s.  In a clamp on a stand.
 TRAINING
GUIDE NOTES
                                                                                                           E21- 31

-------
 WATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                           L21-32
 OPERATING PROCEDURE:
         STEP  SEQUENCE
     INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
F. Sample Treatment
   (Continued)
G.  Sample Analysis
29.  Pipet the solvent layer
    from the neck of the ampul
    (step 22) to the top of
    the mini-column.

30.  Add small amounts of
    benzene to the ampul,
    shake,  allow to separate
    and pipet into column.
                          31. Adjust final  volume to
                              5.0 ml.

                          32. Stopper  the ampul.
29a. Another Pasteur pipet and bulb can be used
     (just the pipet no packing).
29b. Collect as little of the aqueous sodium sulfate
     as possible.

30a. This is a washing step for the ampul, aqueous
     phase and column.  Little to no water should be
     transferred to the column.
30b. The final volume will be 5.0 ml.  Take care not
     to wash with  volumes of benzene which will sur-
     pass this total volume.
                                32a.  This  extract if kept well  stoppered and refrig-
                                     erated,  can be  held 30 days.
 1 .  Check  all  instrument
    parameters.

 2.  Inject 5  ul  of the  stan-
    dard mix  2.

 3.  Determine  if  response  has
    changed.
 4.  Inject  5
    blank.
                                           of method
                           5.  Inject 5  ;jl  of the stan-
                              dard carried  through the
                              procedure.
 la.  They should be the same as used to obtain
     optimum results (section E.I.3).

 2a.  Table 1.
 3a.  Compare results for retention times and response
     (peak height)  with those obtained in E.I.2.

 4a.  No significant peaks should be obtained.
                                 5a.  From  section  F.2.1.Id.

-------
 MATER MONITORING  PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
 OPERATING PROCEDURES
        STEP SEQUENCE
    INFCRMATION/OPIRAIING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
G.  Sample Analysis
   (Continued)
H.  Calculations
   1.  Calculation of
      Results
6. Inject  5 ul of the
   sample/s.
                           7. Compare with standard
                              curve.

                           8. Choose a standard whose
                              concentration will produce
                              a peak that will  approxi-
                              mate that obtained from
                              the sample.

                           9. Calculate the amount of
                              herbicide/s present.
1.  Calculate the micrograms
   of methyl ester per liter
   of sample.
6a.  Inject  5 ul of the duplicate sample.
6b.  Be sure  to time the peaks from the first
     appearance of the solvent peak to the top of
     each peak.

7a.  Step E.3.3,
                                9a.  See the calculation section.
la.  Determine the methyl  ester concentration by
    using the equation below.

    micrograms   A x B x  V
                                                               per  liter
                                                                            Vi*V5
                                                                 where A  = 03L standard    (obtained  in Sample
                                                                           standard  area  Analysis  section,
                                                                                          step G.8)
                                                                       B  = Sample aliquot area (step G.6)

                                                                      Vt  = Volume of total  extract  in micro-
                                                                           liters (the volume to which the
                                                                           extract was concentrated i.e., 5 ml)
                                                                                                           E21-33

-------
HATER MONITORING PROCEDURE:   Determination of Chlorinated Phenoxy Acid Herbicides
                                                                                                          E21- 34
 OPERATING PROCEDURES
                                  STEP SEQUENCE
    INFORMATION/OPERATING  GOALS/SPECIFICATIONS
 TRAINING
GUIDE NOTES
H. Calculations
   (Continued)
   2.  Reporting  Results
                          2.  Convert to niicrogratns of
                             acid per liter of sample.

                          1.  Report results in ing per
                             liter as the acid.
                                                                      V. = Volume of extract injected in
                                                                           microliters

                                                                      V  = Volume of water (sample) extracted
                                                                           in mill liters

                                                                                      Molecular Wt. Acid
                                                                                       Molecular  Wt,  Ester
                                                                 jjg acid = ug ester x


                                                              Molecular Weight:

                                                              2,4-D (acid)   = 222.0
                                                              2, 4-D (ester)  = 236.0
                                                              Silvex (acid)  = 269.5
                                                              Silvex (ester)  = 283.5
la.  Without correction for recovery data.
lb.  Report duplicate and spiked sample results when
    analyzed.

-------
 WATER MONITORING PROCEDURE:  Determination of Chlorinated Phenoxy Acid Herbicides


                               TRAINING GUIDE

 SECTION                           TOPIC

   I*                    Introduction

  II                     Educational Concepts - Mathematics

 III                     Educational Concepts - Science

  IV                     Educational Concepts - Communications

   V*                    Field and Laboratory Equipment

  VI                     Field and Laboratory Reagents

 VII*                    Field and Laboratory Analysis

VIII                     Safety

  IX                     Records and Reports
 *Training Guide materials are presented here under the headings marked*.
 These standardized headings are used through this series of procedures.
                                                                         E21-35

-------
WATER MONITORING PROCEDURES:   Determination of Chlorinated Phenoxy Acid Herbicides
INTRODUCTION
                                                          Section I
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
C.I.la
When the National Interim Primary Drinking Water
Regulations were promulgated they contained the re-
quirement that all Public Water Supplies be moni-
tored for herbicide contamination.  Certain
monitoring frequencies and analytical methods were
prescribed.  A level beyond which public notifica-
tions and other steps were to be carried out was
set and termed the Maximum Contaminant Level (MCL).
The MCL's for the chlorophenoxy herbicides are as
follows:

   2,4, Dichlorophenoxyacetic acid (2,4-D) -
     0.1  mg/liter

   2,4,5-Trichlorophenoxypropionic acid (2,4,5-TP)
     0.01 mg/liter

These materials are used extensively for weed
control in lakes, streams and irrigation canals.
Phenoxy acid herbicides are very potent even at
low concentrations.

-------
MATER MONITORING PROCEDURES:   Determination of Chlorinated Phenoxy Acid Herbicides
 FIELD AND LABORATORY EQUIPMENT
                                                           Section  V
                                 TRAINING GUIDE NOTE
                                                       REFERENCES/RESOURCES
 E.1.3a
 E,3.2b
D.I
D.6.6b
 Because of the many variables inherent in gas
 chromatography, the column packing,  column oven
 temperature and carrier gas flow rate may have  to
 be adjusted to different settings than those given.
 The analyst should strive to reproduce the re-
 tention times  given in the body of the paper as
 guides.  The two things which must be obtained are
 reproducibility and resolution.   When these are
 adequate,  the  system is suitable.

 The analyst must determine the detection  limit for
 each herbicide.   If a sample is  taken through the
 procedure  and  no peaks are obtained,  while peaks
 are obtained for a standard carried  along with
 the sample,  then the analyst is  assured  that
 herbicides  are not present in concentrations above
 his detection  limit.  Whenever reporting  such results
 the analyst should report the detection  limit and
 state  that  no  herbicide is  present above  that
 concentration.   The values  200 ng for 2,4-D and
                             10 pi
 20  ng
 10  M!  for  si 1 vex represent  the published  MCL for
 that compound  in a 5  ml  volume,  that  is the  con-
 centration  factor of  this  procedure.

Use of  high grade  carrier gases are recommended.
However, occasionally  bad cylinders of the Sigon-
Methane gas can  be obtained.   Before attaching it to
the instrument,  a  slight sniff of  the gas should be
taken;  if a "fishy" smell  is  noted, the tank may be
contaminated.  Use of  the gas will produce an off-
scale peak and very noisy base line.    If contami-
nated gas is used  in the instrument, remove  from use
as  soon as detected and replace all traps and purge
with a noncontaminated gas.

OV-210 may be substituted for QF-1.  The OV-210  is a
purified version of the QF-1 and does not tend  to
bleed as much as the QF-1.
                                                                              E21-37

-------
WATER MONITORING PROCEDURES:   Determination of Chlorinated Phenoxy Acid Herbicides
FIELD AND LABORATORY ANALYSIS
                                                          Section VII
                                 TRAINING GUIDE NOTE
                                                      REFERENCES/RESOURCES
VII.5.1
Standards may be prepared from the acids or the
methyl esters.  If prepared from the methyl esters,
the esteriflcation steps are not necessary.  How-
ever, the methyl esters are hard to purchase.

Since samples must be esterified, the standard
should also be esterified.  If anything less than
100% conversion of the acid to the ester 1s ob-
tained in the esterification steps, the standards
will not reflect this and an incorrect analytical
result can be obtained.

Consequently, this procedure has been written using
an esterification step for the standards.
  E21-38

-------
                                  TABLE 2
           RETENTION RATIOS FOR METHYL ESTERS OF SOME CHLORINATED
                  PHENOXY ACID HERBICIDES RELATIVE TO 2,4-D
Liquid Phase
Column Temp.
Argon/Methane
Carrier Flow
Herbicide
2,4-D
Si 1 vex
2,4-D
(minutes absolute)
1.5% OV-17
+
1.95* QF-1
185° C
70 ml/min.
RR
1.00
1.34
2.00
5%
OV-210
185° C
70 ml/min.
RR
1.00
1,22
1.62
All columns glass, 180 cm x 4 mm ID, solid support Gas Chrom Q (100/120 mesh)
                                                                       E21-39

-------
E21-40
               6420
               RETENTION TIME  IN MINUTES

       FIG. 1 Column: 1.5% OV-17+1.95% QF-1,
       Carrier Gas:  Argon (5%) /Methane: 70 ml/min.,
       Column  Temp. 185 C, Defector:  Electron Capture.

-------
 WEDNESDAY, DECEMBER 24, 1975
PART IV;
ENVIRONMENTAL
   PROTECTION
     AGENCY
  WATER PROGRAMS


  National Interim Primary Drinking
     Water Regulations

-------
 .19500
      RULES  AND  REGULATIONS
   Title 40—Protection of Environment
     CHAPTER I—ENVIRONMENTAL
         PROTECTION AGENCY
    SUBCHAPTER  D—WATER  PROGRAMS
              IFBL 464 7)

PART 141—NATIONAL INTERIM PRIMARY
    DRINKING WATER REGULATIONS
  On March 14,  1975, the Environmental
Protection Agency (EPA)  proposed Na-
tional Interim Primary  Drinking Water
Regulations pursuant to  sections 1412,
1414,  1415, and 1450 Of the Public Health
Service  Act ("the Act"), as amended by
the Safe Drinking Water Act ("SDWA,"
Pub. L.  03-523), 40 FE  11990.  EPA held
public hearings on the proposed regula-
tions  in Boston,  Chicago, San Francisco,
and  Washington during the month of
April. Several thousand pages of com-
ments on the proposed regulations were
received and evaluated.  In addition, the
Agency  has received comments  and In-
formation  on  the proposed regulations
from  the National Drinking Water Ad-
visory Council, the Secretary of Health,
Education, and Welfare, and from num-
erous others during meetings with repre-
sentatives ol State agencies, public in-
terest groups and others,
  The regulations deal only  with the
basic  legal  requirements.  Descriptive
material will be provided  in a  guidance
manual for use by public  water systems
and the States.
  The purpose of this preamble to the
flnal regulations Is to summarize the most
significant changes made In the proposed
regulations as a result of  comments re-
ceived and the further  consideration of
available information. A  more  detailed
discussion  of  the comments  and of
changes In  the  proposed  regulations  is
attached as Appendix A.
        WATFR SYSTEMS CovrRr.o
  The Safe Drinking  Water Act applies
to each  "public water system," which  is
denned  Jn Section 1401 <4> of the Act as
"a system for the provision to  the public
of piped water for human consumption.
If such system has at least fifteen service
connections or refrularly serves at least
twenty-five individuals." Privately owned
as well   as publicly owned systems are
covered. Service "to the public"  Is Inter-
preted by EPA to Include factories und
private  housing developments. 
-------
                                             RULES AND REGULATIONS
 These on-site inspections of  water sys-
 tems are more effective in assuring safe
 water to the public than individual tests
 taken in the absence of sanitary surveys.
 The regulations provide that monitor-
 ing frequencies for coliform bacteria can
 be changed by the  entitv with primary
 enforcement  responsibility for an Indi-
 vidual  non-community  system,  and  in
 certain  circumstances for an individual
 community system,  based  on  the  results
 of a sanitary survey.
     MAXIMUM CONTAMINANT LF.VT.IS
   Numerous comments were received by
 EPA on the substances  selected for the
 establishment of maximum contaminant
 levels and on the levels chosen Congress
 anticipated that the initial Interim Pri-
 mary Drinking Water Regulations would
 be based on  the Public Health Service
 Standards  of  1962.  and this Congres-
 sional Intent  has been  followed.  Com-
 ments received on the various levels did
 not contain new data sufficient  to re-
 quire the establishment of  level',  differ-
 ent  from those contained  in  the  Public
 Health Service Standards.
          WATER CONSUMPTION
   The maximum contaminant levels are
 based, directly or indirectly,  on an as-
 sumed consumption of two liters of water
-per  day. The  same assumption was used
 In the 1962 Standards. This assumption
 has been challenged because of instances
 where much  higher  water  consumption
 rates occur. EPA's justification for using
 the  two-liter  figure  is  that,  it already
 represents  an above average water or
 water-based fluid intake. Moreover, while
 the factor of safety may be somewhat re-
 duced when greater  quantities of water
 are ingested, the maximum contaminant
 levels based on the two-liter figure pro-
 vide substantial  protection to virtually
 all consumers. If, as has been suggested,
 a  water  consumption rate of eight liters
•per  day  Is  used  as the  basis  for  maxi-
 mum contaminant level, all of the pro-
 posed MCL's would have to be divided by
 four, greatly  increasing ihe monitoring
 difficulties, and in some cases challeng-
 ing the sensitivity of  accepted analytical
 procedures. It could be expected, in such
 a  ease, that the maximum contaminant
 levels would be exceeded to a  significant
 degree, and that specialized  treatment
 techniques  would be required to  order
 that the contaminant levels would  be re-
 duced. The economic impact  of a move
 In this direction would be enormous.  It
 Is not technically or  economically  feasi-
 ble to base maximum contaminant levels
 on unusually high consumption rates.
            SAFETY FACTORS
   A question  was raised about the fact
 that  different safety factors are con.
 talned in various maximum contaminant
 levels. The levels are  not  intended to
 have a uniform  safety  factor,  at least
 partly because the knowledge  of and the
 nature of the health  risks of the various
 contaminants vary widely. The levels set
 are  the  result of experience,  evaluation
 of the available  data, and professional
 judgment. They have withstood the test
 of time and of professional review, They
 are being subjected to further review by
 the National Academy of Sciences in con-
 nection with development of data for the
 Revised Primary Drinking Water Regu-
 lations.

      MCL's BASED ON TEMPOS'! URE
  A  question  was  also   raised as  to
 whether ranges of maximum contami-
 nant  levels should be established on the
 basis  of  tlie climate in the area served
 by the public water system, as was done
 with  fluoride.  EPA believes that  the use
 of a  temperature scale for fluoride is
 more appropriate  than  for other chemi-
 cals because of the studies available  on
 the   fluoride-temperature  relationship
 and because there is a small margin with
 fluoride  between  beneficial  levels and
 levels that cau.se adverse  health  effects.

            MCL's DELETED
  Three  proposed  maximum contami-
 nant  levels have been eliminated in the
 final  regulations because  they  are  not
 justified  by the avnilable  data.  One of
 these  is   carbon  chloroform   extract
 (CCE>. which  Is discussed  separately
 below. The others are the proposed levels
 for the standard  bacterial plate count
 and  cyanide.  In the  case of the plate
 count, it  is believed  that the coliform
 limits contained in the regulations, com-
 bined with the turbidity maximum  con-
 taminant  level,  adequately  deal  with
 bacterial contamination. However,  EPA
 continues to believe that  the standard
 plate  count  is  a valid  indicator  of
 bacteriological quality of drinking water.
 and recommends that it be used in ap-
 propriate  cases in conjunction with the
 coliform  tests  as an operational tool.
  The proposed maximum contaminant
 level for cyanide was eliminated  because
 the possibility  of cyanide contamination
 can be effectively addressed only by the
 use of emergency action, such as under
 Section 1431 of the Act. EPA's 1969 Com-
 munity  Water  Supply  Study  did not
 reveal a single instance in  which  cyanide
 was present in a water system at a level
 greater than one-thousandth of the level
 at which cyanide is toxic to humans.
  Available data indicate  that  cyanide
 will be present  in water systems  at toxic
 levels only in  the  event of an accident,
 such as a  spill  from  a barge collision.
 Maximum  contaminant levels  are not
 the appropriate vehicle  for dealing with
such rare, accidental contamination.
  Heptachor,    hejstachlor     epoxide
and chlordane  have also  been  removed
 from the list of maximum contaminant
 levels  at least temporarily in view of the
 pending  cancellation   and  suspension
 proceedings under the  Federal Insecti-
cide, Fungicide and Rodenticide  Act in-
 volving those  pesticides. When  the  re-
sults of these proceedings are available,
EPA will again consider whether maxi-
mum  contaminant levels  should  be es-
 tablished for those three pesticides.
         SODIUM AND SUI.FATES
  A number of comments were received
 on the potential health effects of sodium
 and  sulfates.  The  National  Urinking
 Water  Advisory  Council  has  recom-
 mended  that consideration be given to
 the monitoring of these constituents, but
 has not, recommended  the adoption of
 maximum  contaminant  levels  because
 available data do not support the adop-
 tion of any specific levels. EPA has re-
 quested  the  National  Academy of Sci-
 ences  to include sodium and  sulfates
 among  the contaminants  to be  studied
 by NAS, and  to  include information  on
 the health effects of sodium and sulfates
 in  the  report  to be made by  NAS in
 December  1976.
   Since a number of persons suffer from
 diseases  which arc influenced by  dietary
 sodium im.ike and since then: are others
 who wish  to restrict  their sodium in-
 take, it is desirable that the sodium con-
 tent of drinking  water be known. Those
 affected can, bv knowing the sodium con-
 centration in their drinking water, make
 adjustments to their diets or, in extreme
 case?, seek alternative sources  of water
 •a be used for drinking and food prepara-
 tion. I( is recommended that the States
 institute programs [or regular  monitor-
 ing of the sodium content of  drinking
 water served tr>  the public, and  for in-
 forming  {ihvbjfinns and consumers of the
 sodium concentration in drinking water.
  A relatively hifh concentration of sul-
 fate in drinking  water has little or  no
 known laxative effect on regular users of
 the  water,  but   transients using  such
 water  sometimes  experience a laxative
 effect. It is recommended that the States
 institute monitoring programs for stil-
 fp.tss, and  that transients be notified if
 the sulfate content of the water is high.
 Such notification should include  an as-
 sessment of  the  possible physiological
 effects of consumption of  the water
         PCB's  AND ASBESTOS
  An inter a,Ten cy  comment expressed
 concern  for asbestos and PCB's  in the
 environment and noted the need for  at
 least a  monitoring requirement,  if not
 for MCL's. for these contaminants. EPA
 Is also concerned, but for the  moment
 lacks sufficient evidence regarding ana-
 lytical methods, health effects, or occur-
 rence in the  environment  to establish
 MCL's The  Agency is conducting  re-
 search and cooperating in  research proj-
 ects to develop criteria for eslablUhiriK
 needed limits as  quickly  as possible  A
 monitoring study  on a number of organic
 chemical contaminants, including PCB's,
 for which  MCL's are not being  estab-
 lished at this time, will be contained  in
 an organic chemical monitoring regula-
 tion thai Is being promulgated with these
 regulations. Regarding  asbestos,  HEW
 and  EPA are sponsoring  a number  of
studies this year at an approximate cost
 of $16 million to establish  health effects,
 anayltical methods and occurrence
       POIXT or MEASUHEMENT
  Other  comments on  maximum con-
 taminant levels focused on the proposed
requirement that such  levels be  tested
at the consumer's tap. Concern was ex-
 pressed over the  inability of the  public
 water system to control potential sources
                             FEDERAL RfClSTfl, VOL. 40, NO. 241—WEDNESDAY DECEMBER 24,  1775

-------
59568
     RULES AND REGULATIONS
of contaminants  which  are  under the
control of the consumer.
  The promulgated definition of "maxi-
mum contaminant level," | 141,2(d>, re-
tains the requirement that  the  maxi-
mum contaminant level be measured at
the tap except In  the case of turbidity,
which should  be measured at the point
of entry to the distribution system. How-
ever, the definition has  been  expanded
to make  clear that contaminants added
to the water by circumstances under the
control of the consumer  are not the re-
sponsibility  of the supplier of  water,
unless the contaminants result from cor-
rosion of piping and plumbing resulting
from the quality  of the  water supplied.
It should be noted, however, that this
requirement ihould not  be Interpreted
at to discourage  local, aggressive cross
connection control measures.
      ConroHM BACTERIA MCL's

  The  promulgated MCL's for collform
bacteria  are  basically  the 1962  Public
Health Service Standards, with  minor
refinements and clarifications. However,
further changes may be desirable. For
example, the MCL's for  the  membrane
filter analytical method  do not resolve
the question of how many collform bac-
teria are assumed to  be present In  a
single  highly contaminated  sample.
Some laboratories assume an upper limit
of 50, while others Reek to continue to
count Individual  bacteria to a level of
100 or even higher in a single sample
The upper limit assumed will  affect tlie
monthly  average  which is  calculated to
determine compliance  with the MCL's.
  Another question relating to the coli-
form bacteria  MCL's Is the matter of
possible spurious positive samples. As the
regulations are written, all routine sam-
ples taken to determine compliance with
the  MCL's must  be counted,  regardless
of the results of unalysls of any  check
samples  that may be  taken. The reason
for this  is that bacterial contamination
la often intermittent or transient, and as
a result negative  check samples taken  a
day or more after a positive sample can-
not demonstrate that the positive result
was In error. It may be possible, however,
to prescribe a means of dealing with spu-
rious positive  results  without compro-
mising the Integrity of the MCL's.
   A third question concerning the MCL's
for collform bacteria is the relationship
of monthly averages of coliform bacteria
levels to monthly percentages of positive
samples. For  example, the monthly av-
erage  MCL  for   the  membrane filter
method Is violated if the monthly aver-
age exceeds one collform bacterium per
sample. However, for purposes of deter-
mining  whether  the monthly-percent-
age-of-positlve-samples MCL Is violated.
a sample Is counted as positive only If it
contains more than four coliform bac-
teria.  Thus,  It is possible, particularly
when a relatively  small number of sam-
ples Is  taken, for a system  to fail the
monthly average MCL even when no sin-
gle  sample  taken during the month Is
out of compliance with the limit.
   These and other questions concerning
Uie coliform bacteria MCL's will be re-
viewed further by EPA. If review Indi-
cates  that changes  In  the MCL's are
desirable, those changes will be made as
soon as possible but within 8 months, In
time to take  effect at the same  time as
the Initial Interim  Primary  Drinking
Water Regulations.
          ORGANIC CHEMICALS
  The proposed maximum contaminant
levels for organit; pesticides, other than
the three which are the subject  of can-
cellation  and  suspension  proceedings,
have been retained. It Is anticipated that
additional  organic   pesticides  will  be
added to the regulations If surveys of
pesticides In  drinking water being con-
ducted  by EPA  indicate  that  this  Is
needed.
  The proposed  regulations also con-
tained a maximum contaminant level for
organic chemicals obtained by the carbon
chloroform extract (CCE)  method. It
was anticipated by Congress that organic
chemicals would be dealt with  primarily
In the Revised Primary Drinking Water
Regulations because of the paucity of ac-
curate data on the health effects of vari-
ous organic chemicals, the large number
of such chemicals, unccrtHinUles over ap-
propriate treatment techniques, and the
need  for addlttoral Information on the
incidence of  specific  organic chemicals
in drinking water supplies. EPA thought
that the CCE standard might provide an
appropriate means of dealing with or-
ganic chemicals as a class pcndinp action
on the Revised Primary Regulations
  Tlie CCE standard  was originally de-
veloped as u  test for  undesirable tastes
and odors Sn  drinking water. As concern
developed over the health effects of or-
ganic chemicals, the  possibility of using
CCE  as a health standard rather than
an esthetic standard was  considered,
  A« pointed out by numerous comments,
CCE  has many failings as an Indicator
of health effects  of  organic chemicals.
To begin with, the test obtains Informa-
tion  on only  a  fraction  of  the total
amount of organic chemicals in the water
sampled. Furthermore, there Is serious
question as to the reliability of  CCE in
identifying   thosn  organic  chemicals
which are most  suspected of  adverse
health effects. In  addition, there are no
existing  data on  which a specific level
for CCE can be established on a rational
basis. To establish n maximum contami-
nant  level under these  circumstances
would almost certainly do more  harm
than  good. It could give a false sense of
security to persons  served by  systems
which are within the established level
and  a false  sense of alarm to  persons
served by systems which exceed the level.
It also  would  divert  resources  from
efforts  to  find more effective  ways of
dealing  with  the  organic chemicals
problem.
  EPA   believes  that  the  intelligent
approach to the organic chemicals ques-
tion Is to move ahead as rapidly as pos-
sible  along  two  fronts. First,  EPA  Is
adopting simultaneously with these reg-
ulations a Subpart E of Part 141, con-
taining requirements for organic chemi-
cal monitoring pursuant to Sections 1445
and  1450 of the Act.
  The regulation*  require that  desig-
nated public water systems collect sam-
ples of raw and treated water for submis-
sion  to EPA for organlcs analysis. EPA
will analyze the samples for a number of
broad organic pnrameters, Including car-
bon chloroform extract (CCE), volatile
aiid  non-volatile  total  organic  carbon
(VTOC and NVTOO, total organic chlo-
rine  (TOC1), ultraviolet absorbancy, and
fluorescence. In addition, monitoring will
be required for probably 21  specific or-
ganic compounds. Selection  of the spe-
cific  compounds has been based on the
occurrence or likelihood of occurrence In
treated water, toxlclty data and availa-
bility of practical  analytical methods.
Laboratory  analyses will  be  used to
evaluate the extent and nature of organic
chemical  contamination  of drinking
water, to evaluate  the  validity of the
general organic parameters as surrogates
for measures of harmful organic chemi-
cals, and to determine whether there  Is
an adequate basis for establishing maxi-
mum contaminant levels for specific or-
ganies or groups of organlcs.
  Second, EPA is embarking on an inten-
sive  research program  to find answers
to the following four questions:
  1.  What are  the  effects of commonly
occurring organic compounds on human
health?
  2.  What analytical procedures should
be used  to  monitor finished drinking
water to assure that nny Primary Drink-
ing Water Regulations dealing with or-
ganics are met?
  3. Because some of these organic com-
pounds are formed during water treat-
ment, what  changes In  treatment prac-
tices are required to minimize the  for-
mation  of these  compounds in  treated
water?
  4.  What treatment technology  must
be applied to reduce contaminant levels
to concentrations that may be specified
in the Primary Drinking Water Regu-
lations?
  This  research  wiil  involve  health-
effects and  epidemlologkal  studies, In-
vestigations of analytical methodology,
and  pilot plant and flsld studies of or-
ganic  removal  unit  processes.  Some
phases of the research are to be com-
pleted by the  end  of this  year, while
much of the remainder are  to be com-
pleted within  the  next calendar year.
  As  soon  as  sufficient Information  Is
derived  from  the  monitoring  program
and  related research, the Interim  Pri-
mary Drinking Water  Regulations will
be amended so that the organic chemi-
cals  problem can be dealt with without
delay. The  monitoring process will  be
completed within 1 year.
  During the Interim period, while sat-
isfactory MCL's for organic contamina-
tion in drinking water  arc being devel-
oped, EPA will act in specific cases where
appropriate to deal  with organic con-
tamination.  If the EPA monitoring  pro-
gram reveals  serious specific  cases  of
contamination, EPA will work with StRte
and  local  authorities  to  Identify the
source and nature of the problem and to
                            FEDERAL REGISTER  VOl  40, NO. J4J—WEDNESDAY, DECEMBER 34,  1975

-------
                                              iUlES AND REGULATIONS
                                                                         59569
 take remedial action. EPA will also old
 the States in identifying additional com-
 munity  water  supplies  that  require
 analysis.
             PUBLIC NOTICE

   The  public  notice requirements  pro-
 posed In § 141.32 dtd not distinguish be-
 tween community and  non-community
 public water  systems. They  would have
 required that public notice of non-com-
 pllanre  with  applicable  regulations  be
 made by newspaper, in  water bills, and
 by other media for all public water sys-
 tems. These requirements are Inappro-
 priate and Ineffective In the case of most
 non-community  water  systems.  Those
 systems  principally serve transients who
 do not receive water bills from the sys-
 tem and who probably are not exposed
 significantly to the local  media.  A more
 effective approach would be to  require
 notice  that can  inform  the  transient
 before he drinks the system's water, and
 thereby  both  warn  the  transient  and
 provide  an incentive to the supplier of
 water to remedy  the violation. Accord-
 ingly, Section 141.32 as adopted provides
 that in the case of non-community sys-
 tems, the  entity with primary enforce-
 ment  responsibility  shall require  that
 notice be given In a form and manner
 that  will insure that the public using
 the  public  water system  is adequately
 Informed.
  The proposed  public  notice  require-
 ments also  faiLed to distinguish between
 different types of violations of the  In-
 terim Primary Drinking Water Regula-
 tions. Since the urgency and Importance
 of a notice varies according to the nature
 of  the violation  Involved,  § 141.32  as
 promulgated seeks to match the  type of
 notice required with the type of violation
 Involved. Written notice  accompanying
 a  water bill or other direct notice by
 mall is required for all violations  of the
 regulations, including violations of mon-
 itoring requirements, and for the grant
 of a variance  or exemption. In addition,
 notice by newspaper and notification to
 radio and  television stations Is required
 whenever a maximum contaminant level
 Is  exceeded, or wlien the entity with
primary  enforcement  responsibility  re-
 quires such broader notice.
     QUALITY CONTROL AND TESTING
              PROCEDURES
  Section  1401(1)  of the Act  defines
 "primary drinking water regulation"  to
 Include "quality control and testing pro-
 cedures." The promulgated  regulations
 include  testing requirements  for each
 maximum  contaminant level, including
 check  samples and special  samples  in
 appropriate cases. The regulations  also
specify the procedures to be followed  in
 analyzing samples for each of the maxi-
 mum contaminant levels. These proce-
 dures will be updated from time to time
 as advances are made In analytical meth-
 ods. For example, references to "Stand-
 ard  Methods for the Examination  of
 Water and Wastewater" are  to the cur-
rent, 13th, edition, but these references
will  be changed to cit* the HtJh  edition
 when It is available In the near future.
  A key element of  quality control for
 public water systems IE accurate labora-
 tory analysis. Section 14128 of tlie regu-
 lations provides that analyses conducted
 for  the purpose of  determining  com-
 pliance  with  maximum  contaminant
 levels must be conducted by a laboratory-
 approved by the  entity with primary en-
 forcement  responsibility.  EPA  will  de-
 velop as soon as  possible. In cooperation
 with  the  States and  other interested
 parties, criteria and  procedures for lab-
 oratory certification. A  Slate  with pri-
 mary  enforcement  responsibility  will
 have'a laboratory certified by EPA pur-
 suant to the prescribed criteria and pro-
 cedures, and in turn  will certify labora-
 tories within the  State.
  Record-keeping requirements and re-
 ports  to  the State  also  will  assist  in
 quality control efforts.
            RECORD-KEEPING
  Adequate record-keeping Is necessary
 for the proper operation  and administra-
 tion of a public  water system. It  is also
 Important for  providing  information  to
 the  public, providing appropriate data
 for inspection and enforcement activities
 and providing information  on \vhich fu-
 ture regulations  can be  based. Accord-
 ingly, a new § HI.33 has been added to
 the regulations to require that eurh pub-
 lic  water  system maintain records  of
 sample analvses and of actions to r.rrrect
 violations of the Primary Drinking Water
 Regulations.
     ECONOMIC AMD COST ANALYSIS
  A comprehensive economics study has
 been ir.ade of the  Interim Primary Drink-
 Ing Water Regulations. This study esti-
 mates the costs of the regulations, evalu-
 ates the potential economic Impact, and
 considers   possible  material and  labor
 shortages. The results of  this analysis are
 summarized here
  Total Investment costs to coinimmily
 water  systems  to  achieve  compliance
 with these regulations arc  estimated  to
 be between $1,050 and $1,765 million. It
 Is  estimated that non-community sys-
 tems will invest an additional $24 million.
 The range of the estimate ii due to un-
certainty as to the desipn flow thnt will
be used in  installing treatment facilities.
Systems not In compliance will have  to
consider sizing their new components  to
reflect  average daily  flow conditions,  or
maximum  daily now  conditions  in  cases
where system storage is not adequate.
  This  investment will  be  spread over
several  years.  Investor-owned  systems
will bear about one-fourth of these rofts,
 and publicly-owned systems the remain-
 der. It Is not anticipated that, systems will
have difficulty financing these capital re-
quirements.
  In annual terms, national costs are ex-
 pected  to be within the following rungcs:
                            In millions
                            ..  $146 247
                            ..  363 .363
                                 17- 36
Capital costs,.	
Operation:! and mnJntenancy
Monitdrtisg (roxHlne oni>')	
     Total 		  -  M26-54S
  Although  these  aggregate figures are
large, most water consumers will not be
 significantly affected. For those users In
 systems servinE  10,000 persons or  more,
 tl»e average annual  treatment cost per
 capita may increase from less than $1.00
 for systems requiring disinfection and
 lead  control, to  between $15 to $35 for
 control of turbidity and heavy metal re-
 moval.  For systems serving ICEB than 100
 persons, the average  annual per capita
 costs of  disinfection, lead control and
 fluoride/arsenic removal are estimated to
 be between $2.10 and  $11.80. However, if
 turbidity control or heavy metal removal
 were required in' a system of this size
 then costs are expected to range from,
 $52 to  5237 per year  per capita. EPA is
 aware of the serious potential economic
 Impact on  users  in these small systems.
 However, the legislative history specifies
 that  the regulations  should be based on
 costs that can be reasonably afforded by
 large metropolitan or regional systems.
 Further economic  evaluation of  these
 systems is teing  conducted, and realistic
 options for these small systems are being
 reviewed. Options that will be under con-
 sideration include  less costly  treatment
 technologies; formation of  regional sys-
 tems;  and  use  of   alternative   water
 sources. Ind'.is!,rial and commercial  users,
 whether providing their own water or
 using public, systems, are  not expected
 to  be  .MBnifk'jintly  aflor.trd  by  these
 regulations.
  Possible constraints to the implemen-
 tation  of  the interim primary regula-
 tion« were  examined. Although   there
 will be an  increase  in demand for chem-
 icals, manpower, laboratories,  and con-
 struction of treatment facilities, it is not
 anticipated tint any of these factors u-Dl
 be  a  ferious obstacle to  implementation
 of  these regulations  over B reasonable
 time frame.
  For the  reasons giver, above. Chapter
 4(1 of the Code of Federa! Regulations Is
 hereby  amended  by the addition of the
 following new Pan 141. These regula-
 tions will  tnke  effect 18 months  after
 promulgation.
 (II is  hereby crrtifird that the economic and
 Inflationary  lir.pnrts o(  these  regulations
 haTft beer* earefuUf evaluated  in accordance
 with Executive Order  11821J

  Dated:  December 10,  1975

                 RUSSELL E. TRAIN.
                      Administrator
            Subparl A—General
 Sec.
 141.1  App!i'-»!><:ity
 141.2  Definitions
 141.3  Cover;. i;r
 141  4  Variances &nd exemptions
 141 5  Siting rrc|ulrrmei>t«.
 141.8  Effective (intr
   Subpart D—Maximum Contaminant Levels
 141.11 MaXiimiti'.   Gor.iiirjilr.&tit  level** Jor
         Inorganic chenilcals,
141.12 Maximum   cont*mln«i;t  lew:!* for
         organic chemicals.
141.13 Maximum   contaminant  levels for
         turbidity.
141.14 Maximum  microbiological contami-
         nant levels
    Subpart C—Monitoring and Analytical
              Require m« n(i
141 Jl  Microbiological  contamliiaiit   sam-
        pling fuid analytical requirement*.
                             FEDEUM IfOtSTIl, VOL. 40,  NO, 248—WEDNESDAY, DECIMiH 24, 1975

-------
59570
                                             RULES AND REGULATIONS
s*c.
141.22

141.23

141,24

1*1.27
141.28
141 29
       Turbidity  aunpllng  and analytical
        requirements
       Inorganic  chemical lampltng and
        analytical requirements.
       Organic  chemical  sampling  and
        analytical requirements.
       Alternative analytical techniques.
       Approved laboratories.
       Monitoring of  coiistcutKo public
        water »ya terns,

 Subpirt D—Reporting, Public Notification, and
             Hfrcord keeping

141.31  Reporting requirement*.
141.32  Public notification of variances, ex-
        emptions,  and  non-compllanca
        with regulations.
141,33  Becord maintenance.

  AUTHORITY: 6*CB. 1413, 1414, 1445, »nd 1450
of the Public Health Service Act, B8 8tat. I860
(42 U.S.C. 300g-l, 300g-3, 30QJ-4, and 300J-B).
g 141.1
          Subpart A—General

         Applicability,
  This part establishes primary drinking
water  regulations  pursuant  to  section
1412 of the Public Health Service Act, as
amended  by the  Gate  Drinking  Water
Act (Pub. L. 93-523); and related regula-
tions applicable to  public water systems.
§141.2  Definitions.
  As used In this part, the term:
  (a)  "Act" means the  Public Health
Service Act, as  amended by the Safe
Drinking Water Act, Pub. L. 83-523
  Cb)  "Contaminant" means  any  physi-
cal,  chemical, biological,  or radiological
substance or matter in water.
  (c>  "Maximum  contaminant  level"
means the maximum permissible level of
a  contaminant In water which  Is de-
livered to the free flowing outlet  of the
ultimate user of a public water system,
except In the case of turbidity where the
maximum permissible level Is measured
at the point of entry to the distribution
system Cbntaminants added to the water
under  circumstances controlled by the
user, except those  resulting from  corro-
Blon ol piping and plumbing caused by
water  quality,  are excluded  from  this
definition.
   (d>  "Person"  means  an   Individual,
corporation, company, association, part-
nership, State,  municipality,  or Federal
agency.
   (e)  "Public  water system" means a
system for the provision to  the  public
of piped water for human consumption,
II such system has at least fifteen  service
connections or regularly serves  an aver-
age  of at least twenty-five  individuals
dally at least 60 days  out of the year.
Such term -Includes  (1)  any collection,
treatment, storage, and distribution fa-
cilities under control of the operator of
such system and used primarily In con-
nection *ith such system, and <2)  any
collection or pretreatment storage facili-
ties  not under  such control  which are
used primarily  in  connection with such
system. A public water system Is either
a "community water system" or a "non-
community water system."
   (i) "Community water system" means
a public  water system which serves at
least 15 service connections used by year-
round  residents or regularly serves at
least 25 year-round residents.
  (11)  "Non-conununlty water  system"
means a public water system that is not
a community water system.
  (f) "Sanitary survey" means an on-
site  review of  the water source, facili-
ties, equipment, operation and  mainte-
nance of a public water system for the
purpose  of evaluating  the adequacy of
such source, facilities, equipment, op-
eration and maintenance for producing
and  distributing safe drinking water.
  (g)  "Standard  sample"  means the
aliquot of finished drinking water that is
examined for the presence  of coliform
bacteria.
  (h)  "State" means the agency of the
State  government  which  has jurisdic-
tion over public water systems. During
any  period when a State does not have
primary   enforcement   responsibility
pursuant to Section 1413 of the Act, the
term "State" means the Regional Ad-
ministrator, U.S. Environmental Protec-
tion Agency.
    Obtains  all  of its water from, but
Is not owned or operated by, a public wa-
ter  system  to which  such  regulations
apply:
   Does not sell water to  any person;
and
  (d>  Is not  .1  carrier wh'.ch conveys
passengers in interstate commerce.
§ 141.1  Variance1* and  exemption*.
  Variances or exemptions from certain
provisions  of these regulations may be
granted pursuant to Sections 1415 and
1416 ol  the Act by the entity with pri-
mary enforcement responsibility.  Provi-
sions under  Part  142,  Notional Interim
Primary  Drinking  Water  Regulations
Implementation—subpart E  (Variances)
and  subpart  F   (Exemptions)—apply
where  EPA  has  primary  enforcement
responsibility.
§ 141.5  Siting requirements.
  Before a  person may enter into a fi-
nancial  commitment for or Initiate con-
struction of  a new public water system
or  increase the capacity of  an existing
public water system, he shall notify  the
State and,  to  the extent  practicable,
avoid locating part  or  all  of the new or
expanded facility at a site  which:
   (a) Is subject to a significant risk
from earthquakes, floods, flres or other
disasters which could cause a breakdown
of  the public water system or a portion
thereof; or
   (b) EKcept  for  intake  structures, is
within the floodplaln of a 100-year flood
or  Is lower  thnn any recorded high tide
where appvopiiate records exist.
The  U.S.  Environmental  Protection
Agency will not seek to override land use
decisions affecting public water systems
siting which are made at the State or lo-
cal government levels.

§141.6  Effective dale.

  The regulations set forth In this part
shall take effect 18 months after the date
of promulgation.

Subpart B—Maximum Contaminant Levels

§ 141.11  Masimum  contaminant level*
    lor inorganic chemicals,

  (a) The maximum contaminant level
for nitrate Is applicable to both commu-
nity water systems and non-community
water systems. The level! for the other
Inorganic chemicals apply only to com-
munity water systems. Compliance v*ith
maximum  contaminant levels for inor-
ganic chemicals Is calculated pursuant to
i 141.23.
  (b)  The following are  the  maximum
contaminant levels for inorganic chemi-
cals other than fluoride:
                                Levtl,
                             milligram!
Contaminant                   per liter
  Arsenic 	-		  0.08
  Barium	—		  1.
  Cadmium	,	  0.010
  Chromium	,	  0. OS
  Lead  	- —	  O.OB
  Mercury  	.	  0.003
  Nitrate .. .  	
                                                                                M.'HOTO.H	 1?.7U>21.<.	
                                                                                71).7 to 7!< .'1..  ,,    'Jl.5l«l!ft2		
                                                                                75.3 10 !l0.5	    : M-S to 32.!<	
                                    2.0
                                    1.8
                                    16
                                    1,4
 § 141.12   Maximum contaminant  levels
     lor organic clieinirala,

   The following  are the maximum con-
 taminant  levels  for organic chemicals.
 They  apply only  to  community  water
 systems.   Compliance  with  maximum
 contaminant levels for organic chemicals
 is calculated pursuant to 0 141,24,

                               Level.
                             mtlltgTami
                               per liter
 (H)  Chlorinated hydrocarbons:
  Endrln (1.2,3,4.13. lo-hesftchloro-   0.0002
    6.7-epoxjr-l,4.  4»,5,fl,7,8,8s-oeta-
    h;dro-l,4-endo,  endo-5,8 - dl'
    metbano naphthalene).
  Llndana   C 1,2,3.4,6,6-Iiexachloro-   0004
    cyelohiiane, gamma Isomer).
  Methoxychlor    (1,1,1-lVlchloro-   0. 1
            bis  |p melhoxjphenyl)
                                                                                    a,  2 -
                                                                                    ethane) .
                                                                                  Tonnphene
               (ClaHttCl,-Technlo«l  0.005
                                                                                    chlorinated  camphene,  (7-69
                                                                                    percent chlorine).
                             FEDERAL REGISTER VOL.  40, NO. 748—WEDNESDAY,  DECEMBER  14, 1975

-------
                                               RULES AND REGULATIONS
                                  59371
 (b) Chlotopheuouys:
   2,4 - D,  (2,4-Dlehlorophenoxj'aoe-  0- 1
    tic acid).
   2,4,5-TP  Stlvex  (2,4,5-Trichloro-  0.01
    pheaoxypropkmlc acid).

 § 141,13   Maximum contaminant leveli
     fur turbidity.
   The maximum contaminant levels for
 turbidity are applicable to both  commu-
 nity water systems and non-community
 water   systems   using   surface  water
 sources In whole or in part. The maxi-
 mum  contaminant  levels for  turbidity
 In drinking water, measured at  a repre-
 sentative »ntry point(s)  to the distribu-
 tion syste.u, are:
   . as de-
 termined by a monthly average pursuant
 to  I 141.22,  except  that  five or fewer
 turbidity  units may be  allowed  if the
 supplier of water can demonstrate to the
 State  that the higher turbidity does not
 do any of the following:
   (1)  Interfere with disinfection;
   <2>  Prevent maintenance  of an effec-
 tive disinfectant  agent throughout the
 distribution system; or
   (3)   Interfere  with   microbiological
 determinations.
   (b)  Five  turbidity units based  on an
 average for two consecutive  days pursu-
 ant to  5 141.22.
§ 141.14  Maximum niivrukiolugiruT con-
     tuminanl levels.
  The  maximum contaminant levels for
 coliform bacteria, applicable  to com-
munity water  systems  and  non-com-
munity water  systems,  are  as  follows:
   (a>  When the membrane filter tech-
nique  pursuant  to  § 141.21  is used.
the number  of coliform bacteria  shall
not exceed any of the following:
  (1)  One  per  100  miililiters  as  the
arithmetic mean of all samples examined
per month  pursuant to  i 141.21   or
 («;
  (2)  Four per  100  miililiters in more
than one sample when less  than 20 are
examined per month; or
  (3) Four per  100  miililiters in more
than five  percent of the samples when
20 or more are examined per month,
   (bid) When  the  fermentation tube
method and  10 mlllillter standard por-
tions pursuant to I 141.21 (a)  are used.
coliform bar.teria shall not be present in
any of  the following:
  (1)  more than 10  percent  of the por-
tions In any month pursuant to  I 141.21
 (b) or (c);
  (li>  three  or  more  portions  in more
than one sample when less than  20 sam-
ples are examined per month; or
  (iii)  three or  more portions in more
than five  percent of the smnples when
20 or  more samples are examined per
month.
   (2>   When  the  fermentation  tube
method and  100  nitillUter standard por-
 tions pursuant  to § 141.21(a)  are used,
coliform bacteria shall not be present in
 any of the following:
   (i)  more than 60  percent  of 'he por-
 tions in any month pursuant to  § 141.21
  or (c);
   Ul»  five portions  in  more  than  one
sample when  less than flve  samples are
 examined per month, or
   (iii)  flve  portions  in more than 20
percent of the samples when flve or more
samples are examined per month.
   (c) For community  or non-community
systems that ore required to sample  at a
rate of less than 4 per month, compli-
ance with  paragraphs  (a),  (b)(l>, or
(b> (2) of this section shall be based upon
sampling during a 3  month  period, ex-
cept that,  at the discretion of the State,
compliance may be based upon sampling
during  a one-month period.

   Subpart C—Monitoring and Analytical
              Requirements
§ 141.21   Microliiotdfic.il   rontcminanl
     tampling   and   analytical   require-
     ment!.
   (a) Suppliers of water for community
water systems and non-community water
systems  shall analyze for coliform bac-
teria for  the purpose  of determining
compliance with § 141.14,  Analyses shall
be conducted in accordance with the an-
alytical .recommendations  set  forth in
"Standard Methods for the Examination
of Water  and  Wastewater,"  American
Public Health Association, 13th Edition,
pp. 662-688, except that a standard sam-
ple size .shall be employed. The standard
sample used in the membrane fiHer pro-
cedure shall be 100 miililiters. The stand-
ard  sample  uied in  the 5  tube most
probable number *MPNi procedure (fer-
mentation  tube method) shall be 5 times
the standard portion. The standard por-
tion  Is either 10 milliliU'rs or  100 milli-
liter.s as described in $  141.14 tb ' and  'c>.
The  samples shall be taken  at  points
which  are representative  of  the  condi-
tions within  the distribution system.
  <\>1 The supplier of water /or a com-
munity water system shall take coliform
density samples (it  regular  time Inter-
vals, arid in number proportionate to the
population served by  the system. In no
event shall the frequency be less than as
set forth below:

                                  on tit
                                      i
                                      2
                                      3
                                      4
                                      6
                                      6
     ion b^rved:       samples per ?«
25 to  1.000		
1.D01 to 2,500	  .    ......
Z.Mt to 3,300	  ..
3,301104,100,. ...  		
4.:01 t-o 4.90<>		
4.901 to 5,800		,,	
5,801 to £,700-_-	_		     7
0.101 to ;.BOO		  ....     B
7.8'U to 8,500		  ..	     9
8.501 to 0,400  .- .-	--		    10
9.401 to 10,300	.-	    11
10,301 to 11,100..	.-		,    12
11.1«)1 to 12.000		    13
12,01/1 to 12.900.				    14
12,801 to 13.700				    15
13,701 to 54,800	  ....    18
14,001 to 15,500..	    17
15,501 to :a.aoo		    is
16,301 to 17.200	..-    19
17.201 la 18.100   ....	    20
18,101 to 18,800	...	    21
18.901 to IB 800		    22
19,801 to 20,700		    2'J
20,701 to 21.500.		-	    24
21,601 '.023,300	,	    25
22,aOI to 23.200		    29
23201 to 24,000			    27
24,001 to 24,900	.	    28
2»,901 to 25,000	.,„	    2B
25,001 to UB.OOQ		..	    30
                                          28,001 to 33.000...	
                                          33,001 to 37,000			
                                          37,001 to 41,000..		
                                          41.001 to 46,000			......
                                          46001 to 50.000		
                                          5O.001 to 54,000		
                                          64.001 to 58,000	..	
                                          69.001 to 64.000	_		
                                          64.001 to 70.000		
                                          70.001 to 76,000			
                                          76,001 to 83,000	...
                                          83,001 to 90.000		
                                          90,001 W96.000 	,	
                                          86,001 to lll.OOO.		
                                          m.QOl to 130,000	_»	
                                          130.001 to 160,000	,	
                                          160.001 to 190,000		
                                          193.001 la 220.000		
                                          220.001 to 250000	
                                          250.001 to 290.000	..	
                                          203.00! 10 320.000			
                                          320 001 to 380,000		
                                          aeo.ooi to 410,000	...	
                                          410.001 to 450,000			
                                          450.0O1 to 500.000		
                                          500,001 to 550.000	.	
                                          550,001 to 500000..	
                                          803 U01 to 880.000	,	
                                          860,001 to 720.000	
                                          720,001 to 790,000 		
                                          •78D.DO 1 to BIO 000	-	
                                          840,001 to 910.000	
                                          910.UOI to 970.000			-
                                          1)70.0-")! to 1.000 000		
                                          1.050.30!  to 1.140.000	...
                                          1.140301  to 1.230.000		
                                          1.231001  to 1,320000	
                                          1,:!JO.;!01  tol,420.000_	
                                          1.420,001  lo 1. WO 000	
                                          1,020.001  to J .63') 000		---
                                          1.030.001  to t.7X)Oor>_-	 	
                                          1,730.001  l<» 1 Bir'J.SOO		-
                                          1,830.001  '.0 l.UfOO'X)			-
                                          l 370.co«  t--i a.ofio i;fio		
                                         2.0130.0'H  lo 2.37.) COO. 	  	 -  --
                                         2.'J7O,t;Q:  to !.5:0,OOO	-_	
                                         2.MO,001  to !,','50000-	--
                                         2.750.001  tu J.CK'OOnO		
                                         3020.001  to 1.320.000-	,	
                                      35
                                      40
                                      45
                                      60
                                      55
                                      BO
                                      65
                                      70
                                      75
                                      80
                                      as
                                      80
                                      B6
                                     100
                                     no
                                     130
                                     130
                                     140
                                     150
                                     180
                                     170
                                     180
                                     190
                                     300
                                     210
                                     220
                                     230
                                     240
                                     350
                                     280
                                     270
                                     230
                                     290
                                     300
                                     310
                                     320
                                     330
                                     340
                                     350
                                     300
                                     370
                                     3§o
                                     390
                                     400
                                     410
                                     420
                                     430
                                     440
                                     450
                                     480
                                     470
                                     480
                                     480
                                     500

Based on a history of no coliform bac-
terial contamination and on a sanitary
survey by  the  State showing  the water
system to  be  supplied solely by  a pro-
tected ground  water source and  free of
sanitary  defect,?, a community water sys-
tem .--crving 2S  to 1,000 persons,  with
written permission from  the State, may
reduce tills sampling frequency  except
that in no  cose shrill il be reduced to less
than  one  per  quarter,
  (c> The .supplier of water for  a  non-
community water -lysw-m shall .sample  for
coliform  bacteria in each calendar quar-
ter during which  the system provides
water to  the public. Such sampling  shall
begin within two years after the effective
dale of  this part.  If  the Stale,  on the
basis  of  a sanitary survey, determines
that some other frequency is more appro-
priate, that frequency shall be the fre-
quency required under these regulations.
Such  frequency  shall be confirmed  or
changed  on  the  basis  of subsequent
surveys.
  (dM 1 > When the coliform bacteria in a
single sample exceed four per  100 miili-
liters  ), at least two consecu-
tive daily check samples shall be collected
and examined  from the  same sampling
point. Additional check samples shall  be
collected daily, or at a frequency estab-
                                         ,1.320.001 to
                                                    1.K20.000.
                                                    J <«60 000,
                                         3 620,001 U
                                         3 960.C01 to  4.310 fl<».
                                         4 3IUTI01 to  4.090.000.
                                         4.
-------
                                             RULES AND REGULATIONS
 Hshed by the State, until the results ob-
 tained  from  at least  two consecutive
 check samples show less than one con-
 form bacterium per 100 mUlilitera,
   <2>  When collform bacteria occur In
 three or more  10 ml portions of a single
 sample  (| 141.14(b)(l)),  at  least  two
 consecutive dally check samples shall be
 collected and examined from  the game
 sampling point. Additional check samples
 shall be collected dally, or at a frequency
 established by the State, until the results
 obtained from at least  two consecutive
 check samples show no positive tubes,
   (3) When collform bacteria occur In all
 five of  the 100 ml  portions of a  single
 sample  <§ ui.H(b) <2) >,  at  least  two
'dally check  samples shall be collected
 and examined from the same sampling
 point. Additional check samples shall be
 collected daily, or at a frequency estab-
 lished by the State, until the results ob-
 tained  from  at least  two consecutive
 check samples show no positive tubes.
   (4) The location at which  the  check
 samples were  taken pursuant to  para-
 graphs  (d) (1), (2>, or (3)  of this section
 shall not be eliminated from future sam-
 pling without approval of the State. The
 results  from  all collform bacterial analy-
 ses performed  pursuant to this subpart,
 except  those obtained from check sam-
 ples and special purpose samples, shall be
 used to determine compliance with the
 maximum contaminant level for collform
 bacteria as established in ! 141.14. Check
 samples shall not be Included in calculat-
 ing  the total number of samples  taken
 each month  to determine compliance
 with 5 141.21 (b) or   Special  purpose samples, such as
 those taken to determine whether  dis-
 infection practices following pipe place-
 ment, replacement, or repair  have been
 sufficient, shall not be used to determine
 compliance with 5 141.14 or 1141.21 (b)
 or (c).
   
-------
                                              RULES  AND  REGULATIONS
 mental  Protection  Agency,  Office  of
 Technology  Transfer, Washington, D.C.
 20460,1974.
   (2) Barium—Atomic Absorption Meth-
 od,  "Standard Methods for the Exami-
 nation  of Water and  Wastewater." 13th
 Edition, pp,  210-215, or  "Methods for
 Chemical Analysts of Water and Wastes,"
 pp.  97-98,  Environmental  Protection
 Agency, Office of Technology Transfer,
 Washington, D.C. 20460,  1974.
   <3>  Cadmium—Atomic     Absorption
 Method, "Standard Methods for the Ex-
 amination of Water  and Wastewater,"
 13th Edition, pp. 210-215,  or "Methods
 for  Chemical  Analysis  of  Water and
 Wastes," pp,  101-103,   Environmental
 Protection Agency, Office of Technology
 Transfer, Washington, D.C.  20460, 1974.
   (4) Chromium—Atomic    Absorption
 Method, "Standard Methods for the Ex-
 amination of Water  and Wastewater."
 13th Edition, pp.  210-215, or "Methods
 for  Chemical  Analysis  of  Water and
 Wastes," pp.  105-106,  Environmental
 Protection Agency, Office of Technology
 Transfer, Washington, D.C.  20460, 1874.
   (5) Lead—Atomic Absorption Method,
 "Standard Methods  for  the Examina-
 tion of  Water and Wastewater,"  13th
 Edition,  pp.  210-215,  or "Methods  for
 Chemical Analysis of Water and Wastes,"
 pp.  112-113. Environmental  Protection
 Agency,  Office of  Technology Transfer,
 Washington, D.C.  20460, 1974.
   (6) Mercury—Flameless  Atomic  Ab-
 sorption Melhod, 'Methods for Chemical
 Analysis of Water and Wastes." pp. 118-
 126,  Environmental Protection Asenr.y,
 Office of TechnoloRv  Transfer, Wash-
 ington, D.C. 20460,1374.
   <7> Nitrate—Brucine     Colorimetric
 Method.  "Standard Methods for the Ex-
 amination  of Water and  Wastewater,"
 13th Edition, pp.  461-464, or Cadmium
Reduction Method, "Methods for Chemi-
 cal  Analysis  of   Water  and  Wastes,"
 pp.  201-206,  Environmental  Protection
 Agency,  Office  of  Technology  Transfer,
 Washington,  D.C. 20460, 1974.
   (8) Selenium—Atomic     Absorption
 Method, "Methods for Chemical Analysis
of Water and Wastes," p. 145, Environ-
 mental   Protection  Agency,  Office  of
Technology Transfer,  Washington. DC.
 20460.1974.
   (Si Silver—Atomic Absorption Meth-
od.  "Standard Methods  for  the  Ex-
amination of Water and Wastewater",
 13th Edition, pp.  210-215, or "Methods
for Chemical Analysis  of  Water  and
Wastes", p.  146. Environmental Protec-
tion  Agency. Office of Technology Trans-
fer, Washington, DC,  20460,  1974.
   flO>  Fluoride—Electrode      Method,
"Standard Methods for the (examination
of Water and Wastewater", 13th Edition,
 pp. 172-174,  or "Methods for  Chemical
 Analysis of Water and Wastes." pp. 65-
 67.  Environmental  Protection  Agency,
 Office of Technology  Transfer, Wash-
 ington, D.C. 20460, 1974,  or Colorimetric
 Method  with Preliminary  Distillation,
"Standard Methods for the Examination
 of Water and Wastewater," 13th Edition,
 pp. 171 -172 and 174-176, or "Methods for
 Chemical  Analysis   of   Water   and
 Wastes," pp. 59-60, Environmental Pro-
 tection  Agency,  Office of  Technology
 Transfer, Washington, D.C,  20460, 1974,

 § 141.24   Organic  chemical  fanipling
     and analytical r<*i|uSrcmenl9.
   (a)  An analysis of substances for the
 purpose of determining compliance  with
 I 141.12 shall be made as follows:
   (1)  For all community water systems
 utilizing surface water sources, analyses
 shall be completed within one year fol-
 lowing the effective  date of this part.
 Samples analyzed shall be collected  dur-
 ing the period of  the year designated by
 the State as the  period when contami-
 nation by pesticides  is  most likely  to
 occur. These analyses shall  be repeated
 at intervals  specified  by the State but
 in no event less  frequently than ?.t three
 year intervals.
   (2)  For community  water  systems
 utilizing  only  ground  water  sources,
 analyses shall he completed by those sys-
 tems  specified by the  State.
   (b)  If the result of  an analysis made
 pursuant to  paragraph ia)   of this  sec-
 tion indicates that the level  of any con-
 taminant listed  in 5 Hi.12  exceeds the
 maximum  contaminant level, the iu:j-
 plier of water shall  report to the State
 within  7 days and initiate  three addi-
 tional  analysts within one month.
   (c> When the average of lour analyses
 made pursuant to paragraph (b) of  this
 section, rounded  to the .--ame number of
 significant figures as the maximum con-
 canirmui level for the substance in ques-
 tion, exceeds  the maximum contaminant
 level, the supplier of \vntrr shall  report
 to thr State pursuant to 5  Ml.31 and  give
 »ot:re  to the  public pur.sunnt to ? 141.32.
Monitor):)!: after public notification shall
be at a frequency designated by the State
 and  slinH CL.-nUr.vie unt:l  the maximum
 contaminant  level has not been exceeded
 in two surre'P-ive .iimpU's  o:  '.ailil  n
 monitoring «'hc'.iule as a  condition  to n
Viirjrtiu'o, exemption or enf'irrf-inoii: ;••<•-
 tion shall become elfeet:vc>
     For the initial  anal VMS required
by paragraph ;a>  il>  and  (2- of  this
section, data for  surface water acquired
within  or.o year  prior  to the effective
date of this  part and ti:ua  lor ground
water acquired within  three  yen:s pn<>;
ta the  effec.tivtt date of this uu:*. r:;.sy be
substi'.ulrd a; the ai'-er<'Kin of (he  St'.ite.
    Analyses made to determine com-
pliance with  3141.l2-
chtorine Pesticides in  Industrial  Efflu-
ents,"  MDQAKL, Environment!  Pro-
tection Agency. Cincinnati. Ohio, Novem-
ber 28. 1973.
  (ft  Analyses made to iiel•<>'cms, the State may modify the moni-
 •or::)B  requirements  imposed  by  this
 i:?rt w the extent that the interconnec-
 :»n of the syseirc.s junfirs treating them
 c.<  a .'••ingle  system for monitoring pur-
 J;U-P« Any modified monitoring shall be
 conducted pursuant lo a schedule  speci-
 fied by the SUite a::d concurred in by the
 Admini'-trator of the U.S  Environmental
 Protection Agency,
 Cubpart D—Reporting, Public Notification
          and Record Keeping
 5: 111.31   Kfpnriiiic icijui
   'ai Exrept  uhi'i'i* a .•.•lurti-r
 period  is  specified   in  this  part,  the
 supplier ni water shall rfport to the State
 uitlur, 40 days fdknuns; a test, measure-
 ment or acaiy-15 rinuimi. to be made by
 ihS part, the  rc-u!t. of that test, nieas-
 ur'";ncnt or a:i:ily-it.
   ' b> Th" st!|.>: I lor <»f v.atf r .s;ial! report
 to the h'tate u itiiin  :!i houiv-i '.he failure
 to comply v,:!h  any primary di'i/ukir.K
 v.:>tcr  n'i;ul:i'.i tti  ' incl jdir.g  failure  to
 co:n|i7y  ui'Ji miinitorir.g  reqairfmenl.-;!
 sc' :orth in llii-; iv.rt.
   '(•; Ti'.e suppik-r  c-f  '.va'er  is not ve-
 ritu'vtl 'o  rcpori iHiulvl'.i ;U results to the
 SI.;;!" in ca=c-s wlicrt' a  State laboratory
;.".vrl.')!'i::s  the  •.u:a!y:-'i--i  and reports  the
 rc-.ults to the State n;T;re wliich would
 nnirially  i-fet:n o .--.m-V. notiiication irom
 tin- .'".ipplier.

!j JH..12   I'llllilV I>rltrable  maximum
<.--.>m.imi:iant level cstal-lished in Subpavt
D, fruls to  comply  with  an applicablR
ti-.itiiig procedure eylabii^hcd In Subpart
C of this  part, is granted a variance or
an exemption  from an applicable maxi-
mum contaminant levol, fails  to comply
with the  requirements  o:'  any schedule
prescribed pursuant  to a variance or ex-
 cmiJlion, or fails U>  perform any moni-
toring required pui'Miant to Section 1445
 >'a> of the Act, the supplier of water shall
notify persons served by  the  system  of
 the failure or grant by Inclusion of a no-
 tice in the first set of water bills of the
system issued  after the failure or  grant
                            FEDERAL HIGISIIR, vot.  40, NO. 24i—WEDNESDAY,  DECEMBER  1*1, i»?s

-------
59574
      •ULES  AND  REGULATIONS
and In any event by written notice within
three months. Such  notice shall be  re-
peated at least once every three months
so long aa the system's failure continues
or the variance or exemption remains ID
effect. If the system issues water bills less
frequently  than quarterly, or  doe's  not
Issue water bills, the notice shall be made
by or supplemented by another form of
direct mall.
  (b) If a  community water  system has
failed to comply with an applicable mai-
Imunt contaminant level, the supplier of
water shall notify the public of such fail-
ure.  In  addition to the notification  re-
quired by paragraph (a;  of this section,
as follows:
  (1) By publication on  not less than
three consecutive days In a newspaper or
newspapers of general circulation in  the
area served by the  system. Such notice
shall be completed within fourteen days
after the  supplier  of  water leams  of
the failure,
  (2) By furnishing a copy of the notice
to the radio and television stations serv-
ing the area  served by the system. Such
notice shall  be furnished within seven
days after the supplier of water learns
of the failure,
  (c) If the area served by a  community
water system Is  not served  by a dally
newspaper of general circulation, notifi-
cation by newspaper required  by para-
graph (b) of this  section shall Instead be
given by publication on three  consecutive
weeks in a weekly newspaper of general
circulation serving the area. If no weekly
or dally newspaper  of  general circula-
tion serves the area, notice shall be given
by posting  the notice In post offices with-
in the area served by the system,
   (d) If a non-community  water sys-
tem fails to comply with an applicable
maximum  contaminant level established
in Subpart B ot this pert  fails to comply
with an  applicable  testing  procedure
established In Subpart C  of this part, la
granted a variance or an exemption from.
an  applicable maximum contaminant
level, fails to comply with the require-
ment of any schedule  prescribed pursu-
ant  to a variance or exemption or falls to
perform any monitoring required pursu-
ant  to  Section 1445'a)  of the  Act.  the
supplier of water shall  given  notice of
such failure or  grant  to the persons
served by the system. The form and man-
ner  of such notice shall be prescribed by
the  State, and  shall  Insure  that  the
public using the system Is adequately In-
formed of  the failure or grant.
   (e) Notices given pursuant to this  sec-
tion shall be written In a manner reason-
ably designed to inform  fully  the users
of the system. The  notice shall be con-
spicuous and shall rot use unduly tech-
nical language,  unduly small  print or
other methods which would frustrate the
purpose of the notice.  The  notice shall
disclose all material facts regarding the
subject Including the nature of  the prob-
lem and, when appropriate, a clear state-
ment that  a  primary  drinking  water
regulation has been violated and any pre-
ventive measures that should b« taken by
the  public. Where appropriate, or where
designated by the State, bilingual notice
shall be given. Notices may include a  bal-
anced explanation of the significance or
seriousness to the public  health of  the
subject of (be notice, a fair explanation
of steps taken by  the system to correct
any problem and the remits of any addi-
tional sampling.
   (f) Notice to the public required by
this section may be given by the State on
behalf of the supplier of water.
   (g) In any Instance In which notifica-
tion by mall la required by paragraph (a)
of this section but notification by news-
paper or to radio or  television stations
la not required by paragraph (b) of this
section, the State may order the supplier
of water to provide notification  by news-
paper and to radio and television stations
when circumstances make more immedi-
ate  or  broader  notice appropriate to
protect  the public health.
§ 141.33  Record maintenance.
   Any owner or  operator  of  a punlic
water system subject to the provisions of
this  part shall retain on Its premises or
at a convenient location near Its prem-
ises the following records:
   (a) Records of bacteriological analyses
made pursuant to this part shall be kept
for not less  than 6  years. Records of
chemical analyses made pursuant to this
part shall be kept for not less than 10
years. Actual laboratory reports may be
kept, or data may be transferred to tab-
ular summaries,  provided  that the fol-
lowing  information Is  included:
   (1) The date, place, and time of sam-
pling, and the name  of the person  who
collected the sample;
   (2> Identification of the sample as to
whether it was  a routine distribution
system  sample, check  sample, raw or
process water  sample or  other special
purpose sample;
   (3) Date of analysis;
   (4) Laboratory and person responsible
for performing analysis;
   (5) The analytical technique/method
used; and
   <6> The results of  the analysis.
   (b) Records of action  taken by the
system  to correct violations of primary
drinking water regulations shall be kept
for a period not  less  than 3 years  after
the last action taken  with respect to the
particular violation involved.
   
-------
                                    TECHNICAL REPORT DATA
                             (Please read Instructions on th? rererse before completing]
 1. REPORT NO.
  h'A-430/1-80-006
 4. * TLE ANDSUBTITLE
Methods  for  the Determination of  Chemical  Contaminants
in  Drinking  Water Training Manual
|3. RECIPIENT'S ACCESSIOf*NO.
!          Fee?'   &Q31
jS~ REPORT DATi
   April  1980	
                                                            6. PERFORMING ORGANIZATION CODE
 . ALTHCRSSl

  John  D,  Pfaff
                                                            8. PERFORMING ORGANIZATION REPORT NO.
9. PEBFORV.!NG ORGANIZATION (SiAME AND ADDRESS
                                                             1O. PROGRAM ELEMENT NO.
Environmental  Protection Agency
Office of  Water Program Operations
National Training & Operational Technology  Center
Cincinnati,  Ohio 45268
                                                             1 1. CONTRACT/GRANT NO,
 12. SPONSORING AGENCY NAME AND ADDRESS

  Same
                                                             13. TYPE OF REPORT AND PERIOD COVERED
                                                             14. SPONSORING AGENCY CODE
 15. SUPPLEMENTARY NOTES
  Supercedes  P8  284392/AS
 16. ABSTRACT

  This laboratory  manual  is designed to contain  analytical procedures  for  all  parameters
  listed in  the  National  Interim Primary  Drinking Water Regulations  (NIPDWR).   Some
  procedures  ray be  carried out by operators  or  laboratory technicians  with  little or
  no experience  (chlorine, trubidity,  fluoride,  nitrate) while other parameters require
  understanding  and  experience in using sophisticated analytical equipment (atomic
  absorption,  gas  chromatography).  Parameters  included are procedures  for silver,
  cadmium, chromium,  lead, mercury, arsenic,  selenium, nitrate, fluoride,  barium,
  chlorine,  turbidity,  perticides, herbicides.   It is written in a step-by-step format.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
  Analysis, Chemical  Analysis, Chlorine,
  Turbidity, Metals,  Silver, Cadmium
  Chromium Fluoride,  Barium, Pesticides,
  Herbicides, Potable Water.
                                               b.lDENTlFlERS/OPSN ENDED TERMS
                                                                           COSATI Field/Group
                                                                           07B,  07C,  14B
 3. D:S-RIBUT!ON STATEMENT
Release to the  public
                                               19. SECURITY CLASS (This Report!
                                                   Unclassifi ed
                                                                          21, MC. OF PAGES
                                               20. SECURITY CLASS (This pate)
                                                   unclassi tied
                                                                          22. PRICE
EPA Form 2220-1 (9-73)

-------