QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Pagel of 3 6
Releasable Asbestos Field Sampler
         (RAFS)
     Operation Manual
        I   •  " *?
         ft! JB A — 4
              HRTI
              INTERNATIONAL
       For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 2 of 3 6
Releasable Asbestos Field  Sampler
                  (RAFS)
           Operation Manual
        Environmental Quality Management, Inc.
              1800 Carillon Boulevard
               Cincinnati, OH 45240
         U.S. Environmental Protection Agency
           26 W. Martin Luther King Drive
               Cincinnati, OH 45268
                RTI International
               3040 Cornwallis Road
          Research Triangle Park, NC 27709
                September 16,2011
              For US EPA Internal Use Only

-------
  QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page3 of 3 6
                            Table of Contents

1   RAFS Overview	5
  1.1   Description	5
  1.2   Theory of Operation	5
  1.3   Safety	10
2   Unpacking Instructions	14
  2.1   Unpacking	14
3   RAFS Setup	18
  3.1   Hardware Inspection                                                    18
  3.2   Connect Auxiliary Components	19
4   Operation and Sampling Conditions	20
  4.1   Parameters and Values	20
  4.2   Operating Settings	21
  4.3   Test Location Preparation	22
  4.4   Aerosol Concentration/Sample Volume	26
  4.5   Loading/Unloading Filter Cassettes                                       28
5   Step-By-Step RAFS Operation                                               29
6   Maintenance, Quality Control, and Calibration	31
  6.1   Cleaning the RAFS                                                     31
  6.2   Quality Control	31
  6.3   Equipment Performance Check and Calibration	31
7   Troubleshooting	33
8   References	34
Appendix A: Example Forms	35
                           For US EPA Internal Use Only

-------
  QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 4 of 36
                                     Figures


Figure 1. Schematic of theRAFS	6
Figure 2. RAFS at Field Evaluation Site	7
Figure 3. Air flow diffuser at tunnel inlet and filter cassette mounting assembly at tunnel outlet. 7
Figure 4. Side view of RAFS showing traverse path of raking mechanism	8
Figure 5. Filter cassette mounting holder and optical particle counter platform at tunnel exit	9
Figure 6. RAFS components in packing cases	14
Figure 7. Photographs of RAFS auxiliary equipment listed in Table 3. (From left to right,
primary air flow calibrator, vane anemometer, gasoline generator, soil moisture meter, optical
particle counter, and meteorological station)	16
Figure 8. Aluminum (or equivalent non-porous surface) for collection quality assurance
equipment blank	17
Figure 9. RAFS electrical connections and GFCI. Note: Location of in-line critical flow orifice.
	18
Figure 10. RAFS control panel	21
Figure 11. Uneven surface beneath tunnel wall must be relatively even to eliminate the release
of air from beneath the frame	23
Figure 12. Air gap beneath tunnel wall  is sealed with soil	23
Figure 13. RAFS operating in soil  with vegetation	24
Figure 14. Proper contact of rake tines with soil resulting in grooves (~!/2-inch depth)	25
Figure 15. Photographs of 25 mm dia. 0.8 jim pore MCE filters to assess proper loading of
particles for direct-transfer TEM analysis, a) unused filter, b) lightly loaded filter, c) ideal,
properly loaded filter, d) overloaded filter, e) very overloaded filter	28


                                      Tables
Table 1. Main Components of the RAFS                                              15
Table 2. RAFS Auxiliary Equipment                                                 15
Table 3. Sampling Filters and Support Materials                                      16
Table 4. Recommended Operating Settings                                           20
Table 5. Guidance to Determine Target Sampling Time based on RAFS Generated
Particulate Number Concentration and Sample Flow of 13.5 Lpm	27
Table 6. Troubleshooting for RAFS                                                  33
                             For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 5 of 3 6
                           1     RAFS  Overview
1.1   Description

       The Releasable Asbestos Field Sampler (RAFS)1 is a field instrument that provides an in-
situ measurement of asbestos releasability from consistent and reproducible mechanical agitation
of the source material such as soil. The RAFS was designed to measure concentration (asbestos
structures per cubic centimeter of air) and an emission rate (asbestos structures per second) from
soil. The RAFS operates in situ under actual soil conditions with representative moisture content
and grain size.  Kominsky et  al. (2010) describes the development and field validation of the
RAFS.(1)

       RAFS utilizes a  raking motion to provide the energy that  releases  particulate material
from the soil and aerosolizes the  asbestos fibers.  A gentle airflow transports the generated
aerosol  laterally inside  a tunnel to  one  end  where filter sampling  cassettes or  real-time
instruments are used to measure asbestos and particulate release.

       The RAFS operates as a semi-autonomous instrument with operator-controlled set points
such as speed of the raking mechanism and air velocity through the wind tunnel. These set points
can be adjusted and optimized by the operator for use in a variety of  conditions.


1.2   Theory of Operation

       The RAFS consists of a variable-speed HEPA-filtered fan attached to a tunnel  (6-inches
by 6-inches by 24-inches) with an open bottom for exposure to the test matrix soil (Figures 1 and
2). The fan discharges the air at the tunnel inlet through diffusers  to  evenly distribute the airflow
(Figure 3). A variable-speed motorized rake mechanism inside the tunnel provides consistent and
reproducible agitation of the top 1A inch of soil (Figure 4). The rake  mechanism has 10 tines that
oscillate  slightly as it traverses  the tunnel  back and forth to agitate the soil to aerosolize the
asbestos fibers. An attachment at the tunnel exit can support up to three 25-mm-diameter mixed-
cellulose ester (MCE) membrane filter cassettes with 50-mm electrically conductive extension
cowls for asbestos collection and analysis using direct-transfer transmission electron microscopy
(TEM), as well as serves as a platform for positioning an optical particle counter (Figure 5). This
aspect of the RAFS design permits collection of concurrent filter samples for different sampling
periods with resultant varied air volumes to ensure that a sample is obtained with an acceptable
particulate loading for analysis  using direct-transfer  TEM.  A typical sampling period ranges
from 10 to 60 minutes, depending on the filter particulate loading.
    U.S. Patent No. 7,758, 812 (July 20, 2010), Environmental Quality Management, Inc.,
    1800 Carillon Boulevard, Cincinnati, OH 45240.
                              For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 6 of 36
       Each filter assembly is attached with rubber and copper tubing to an electric-powered
(110-volt  alternating  current)  1/10-hp vacuum  pump operating  at  an  airflow rate  of
approximately 13.5 liters per minute (1pm). Each pump is equipped with a flow-control critical
orifice that maintains the flow rate at approximately 13.5 1pm.

       The RAFS collects anisokinetic samples where the freestream velocity is greater than the
sample velocity.  Anisokinetic sampling does not introduce  a positive bias to  the  measured
asbestos concentration because the Stokes number for the aerosolized asbestos fibers is less than
0.01.  Under  these  conditions, inertia  effects  are  negligible and the freestream to  sample
concentration ratio is unity.(2)
                                                                                       1
Figure 1. Schematic of the RAFS.
                              For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 7 of 3 6
             Figure 2. RAFS at Field Evaluation Site.
Figure 3. Air flow diffuser at tunnel inlet and filter cassette-mounting
                    assembly at tunnel outlet.
                  For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 8 of 3 6
 T®T
Figure 4. Side view of RAFS showing traverse path of raking mechanism.
                   For US EPA Internal Use Only

-------
QLOG#
S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 9 of 3 6

Figure 5. Filter cassette mounting holder and optical particle
              counter platform at tunnel exit.
               For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 10 of 36
1.3   Safety

       Inhalation of any type of asbestos will  increase the risk of lung cancer, mesothelioma,
and non-malignant lung and pleural disorders including asbestosis,  pleural plaques, pleural
thickening, and pleural effusions. Exercise sufficient precautions to avoid inhalation exposures
when handling materials suspected of containing asbestos such as soil.  Avoid generating dust
during cleaning of the wind  tunnel.  Use of pressurized air duster or dry wiping may  generate
airborne asbestos fibers.  The surfaces should be damp wiped using a lint free cloth moistened
with a standard all-purpose surface cleaner. Exercise sufficient precautions to prevent  potential
electrical hazards associated the 110-VAC power cords. Operators should strictly follow the site-
specific Health and Safety Plan. If a Health and  Safety Plan is not available, consult an American
Board of Industrial  Hygiene (ABIH)  certified industrial  hygienist to identify  the  personal
protective equipment that should be used.
                              For US EPA Internal Use Only

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 14 of 36
                   2     Unpacking Instructions
2.1   Unpacking

      Before removing any RAFS components from the  carrying cases  (Figure  6), the
arrangement of the various items should be noted,  so that repacking can be done easily. To
prevent damage to the RAFS, use the handles at each end of the RAFS main body when carrying
the instrument. The main body is constructed of lightweight aluminum with clear-plastic side
panels and is susceptible to damage.
                   Figure 6. RAFS components in packing cases.
                                     14

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 15 of 36
2.2   Packing Inventory

       Table  1  lists the main components of the RAFS. Refer to Figures  1  and 2 for an
explanation and diagram of each item in Table 1. Table 2 lists the RAFS auxiliary equipment.
Figure 7 shows examples of the auxiliary  equipment. Other manufacturers produce equivalent
equipment; the user should decide what manufacturer and instrument model  is best for their
needs. Of the equipment listed  in Table 2, only the meteorological station and the optical particle
counter are considered optional. The meteorological station and particle counter are not required
if the RAFS is to be used independently;  if RAFS releasability  data are to be compared with
activity-based sampling data, the both instruments  are highly recommended.  Table 3 lists the
filters and the specific accessories required to run each test. Example data collection forms are
located in Appendix A. The user should create labels and forms to suit their specific needs.

	Table 1. Main Components of the RAFS (see Figures 1 and 2)	
  Quantity
                               Description
     1

     1

     1
     3

     3

     3

     1

     1
RAFS assembly (control panel, wind tunnel, raking mechanism, air flow diffuser,
and filter cassette holder and platform).
Axial flow fan and 90° supply air duct.

HEPA filter with rubber connection assembly.
110-VAC rotary vane, pulsation-free oiless vacuum pump (1.1 cfm free air flow,
20" Hg max vacuum, 10 psi max pressure.
1.4 mm orifices ("A") to provide 13.5 Lpm at altitudes less than 2650 feet.
1.5 mm orifices ("B") to provide 13.5 Lpm at altitudes between 2650 and 6900
feet.
Wind direction indicator.
Wind Screen.
                          Table 2. RAFS Auxiliary Equipment
   Quantity
                               Description
             Primary standard air flow meter; flow range 0-20 Lpm.
             Thermal anemometer (30 - 1,000 fpm) or rotating vane anemometer (30 -
             1,000 fpm).
             Hand-held optical particle counter.

             Soil Moisture Meter.
             Aluminum (non-porous equivalent surface) approximately 1-ft by 3-ft for
             collection of equipment blank quality control sample (see Figure 8).
             110-VAC power supply or (portable generator, 1000W, minimum).
             Hand-held meteorological station to measure temperature, wind speed, %
             relative humidity, and barometric pressure (optional).
                                          15

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 16 of 36
             Hand-held optical particle counter (optional).
             	Table 3. Sampling Filters and Support Materials.1
     Quantity
                                    Description
                25 mm diameter, O.Sum pore size mixed cellulose ester (MCE) membrane
                filters with 50 mm electrically-conductive extension cowl to reduce static
                charge.
                Filter labels with unique sample identification code.

                Field sampling and instrument data collection form.

                Chain-of-custody form.
                Lint free wipes and surface cleaner in spray bottle for RAFS
                decontamination.
3

1
1

1
          This listed materials are those needed for a single test. These materials are provided by the
          RAFS user.
Figure 7. Photographs of RAFS auxiliary equipment listed in Table 3. (From left to right,
primary air flow calibrator, vane anemometer, gasoline  generator,  soil moisture meter,
optical particle counter, and meteorological station).
                                           16

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 17 of 36
Figure 8. Aluminum (or equivalent non-porous surface) for collection quality assurance
                                equipment blank.
                                       17

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 18 of 36
                            3    RAFS Setup
3.1  Hardware Inspection

Electrical Connections

      Power connections for the blower and the agitator mechanism are on the RAFS main
body (Figure 9). The vacuum pump power connections are wired into the pump platform that is
connected to the RAFS main body via a standard plug. Inspect these connections making sure the
fittings are snug before applying power to the system. A GFIC switch is located on the control
panel (Figure 9).
                                                  Location of in-line
                                                  critical flow orifice
                   Figure 9. RAFS electrical connections and GFCI.
                     Note: Location of in-line critical flow orifice.
Rake Mechanism

      Figures 1 and 9 show an illustration of the rake mechanism. With the RAFS slightly tilted
upward,  observe the rake travel through a full cycle to determine that it is working properly.
There should be free and consistent movement of the rake mechanism. Caution: Obstruction of
rake mechanism movement can overload the electrical circuit causing the fuse to blow.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 19 of 36
HEPA Filter and Axial Flow Fan Assembly

      Fasten the axial flow fan and 90° air supply duct with air diffuser grate assembly to the
RAFS. Inspect the fan to ensure there is free movement of fan blades. Securely fasten each of the
six fasteners (Figure 9) to ensure an airtight connection.

3.2  Connect Auxiliary Components

Vacuum Pumps

      Position the platform with the three vacuum pumps on the top of the RAFS main body
(Figure 9). Secure the pump platform with the latches (Figure 9). Connect the power cord to the
electrical outlet box  on the RAFS (Figure 9). Note: Each pump has an initial startup power
requirement of approximately 4-5 amps. The vacuum pumps should be started in sequence to
minimize the possibility of over loading a 15 amp power supply circuit. During operation, the
power requirement is reduced to approximately 4.5  amps (1.5 amps per pump). Install the in-line
critical flow control orifice for each pump (Figure 9).

Filter Sampling Cassettes

      Install the sampling table at the outlet of the RAFS (Figures 5 and 8). Slide the hooks at
the  end  of the sampling table over the two pins  located immediately inside the RAFS wind
tunnel. The pins and latches are designed to insert the sample cassettes the proper distance inside
the RAFS wind tunnel.  The open-face, 25 mm diameter, 0.8  |im pore size, mixed-cellulose ester
(MCE) membrane filters in cassettes with  a 50 mm extension cowl are positioned in the three
rubber grommets at the end  of the sampling table. The tension provided by the grommets
properly aligns the cassettes to the airflow inside  the tunnel (Figure 5). Typically, three filter
cassettes  are mounted concurrently. However, the user can use only one or two cassettes, if
desired.

3.3 Power Supply

      The RAFS can  operate from either 110-VAC line power or from a gasoline-powered
portable generator (>15 amps). The generator should not be positioned closer than 50 ft from the
operating RAFS unit (Figure 7).

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 20 of 3 6
           4     Operation and  Sampling Conditions
4.1  Parameters and Values

      Table 4 lists the  required operating parameters and values for proper operation of the
RAFS. Typical operation has the fan operating at maximum speed. The user can decrease the fan
speed to increase the sample collection time. This step is  desired if additional  air volume  is
needed to achieve the desired analytical sensitivity. The rake speed is typically operated at full
power. The user can decrease the rake speed when operating in sandy, loose soil to increase
sample collection time to provide additional air volume to increase analytical sensitivity.
Table 4. Recommended Operating Settings
Item
(Air Velocity)
(Traverse Speed)
Sampling Pump
(Flow Rate)
Target Setting
280-380 fpm
1 cyc/20-30 sec
13 - 14Lpm
Adjustment Location
RAFS Control Panel
RAFS Control Panel
Fixed Value
      Figures 9 and 10 show the RAFS control panel.  The blower (wind) and agitator (rake)
mechanisms are powered on and adjusted (slow to fast) independently. Bulb lights provide a
visual signal the respective systems are powered-on.

      The flow control orifice installed in-line with each pump fixes the  sampling flow at
approximately 13.5 Lpm. It is recommended the user confirm the actual  sample flow rate with a
primary air flow calibrator (Figure 7).

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 21 of 36
                            Figure 10. RAFS control panel.
4.2   Operating Settings

Air Velocity

       The air velocity  through the RAFS tunnel must be measured both before and after
sampling event. The measurement before sample collection is necessary to determine the air
velocity is within the specified operating range (Table 4). Note: Since variations in the test
matrix surface characteristics may influence the air velocity exiting the tunnel, this measurement
should be on a surface that is representative of the test location.

      Place the RAFS on the test surface with the tunnel exit facing downwind; i.e., the wind is
not blowing into the tunnel exit. Adjust the fan control knob to full  speed and then turn on both
the fan and rake agitator.

       The air speed through the RAFS is measured using a calibrated vane anemometer (e.g.,
Davis Anemometer  4-inch or equivalent) or a thermal  anemometer (e.g. TSI Model 9535 or

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 22 of 3 6
equivalent). The instrument must be calibrated against a NIST-traceable standard and operated in
accordance with the operating manual. The air velocity measurements are made at approximately
2 inches inside the exit face of the wind tunnel. The average air velocity is reported as feet per
minute.

       Vane Anemometer—The vane anemometer is positioned inside the tunnel and air flow is
measured for exactly one minute. This yields average air velocity in the tunnel over the 1 minute
period.

       Thermal anemometer—The  instrument probe is inserted through  each of the three
openings at the tunnel exit (Figure 9 and 10). Three equal-distance velocity measurements are
made at each vertical height yielding nine measurements. (Note: There should be no movement
of the probe during the measurement period (ACGIH Ventilation Manual, 2005). To determine
the velocity a straight numerical average of nine individual point velocities is  calculated.
Rake Speed Cycle

       One full cycle consists of two lengths of the tunnel; i.e., back and forth once. The test
should begin with the rake mechanism at the tunnel outlet and the control knob set at full power.
The counter should be reset to zero before beginning each test.  An automatic counter connected
to the mechanism is mounted on the control panel (Figure 10). The RAFS rake speed (time for
one full cycle) is measured with a stopwatch. The rake speed is the traverse length (36 inches per
full cycle) divided by the measured time.

       Unobstructed, the rake will complete one full  cycle approximately every 20 seconds.
However, the rake speed  may be slower due to factors such as  depth and the physical
characteristics of the soil (i.e., moisture, porosity,  grain size,  and presence of vegetation and
aggregate).
4.3   Test Location Preparation

       Preparation of the test location for RAFS sample collection is required. The terrain must
be relatively even (e.g., flat) and be approximately 1 ft wide and 3 ft long so the RAFS tunnel
frame base maintains solid contact with the soil. The RAFS tunnel has a footprint of 6-inches
wide by 24-inches length.  The RAFS tunnel edges should firmly contact the soil. An uneven
surface beneath the tunnel walls must be evened to prevent the release of air beneath the walls
(Figure 11). Small gaps between the RAFS walls and the soil can be filled with extra soil (Figure
12).

       Level  the surface beneath the tunnel walls to ensure the rake tines can reach  a depth of
between H-inch to 3/4-inch. The rake tines must maintain consistent contact with the soil along
the entire length of traverse during sample collection. Failure to maintain consistent contact of
the rake tines  with the soil will reduce the quantity of asbestos fibers aerosolized from the soil.

-------
QLOG#
S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 23
of 36

      Figure 11. Uneven surface beneath tunnel wall must be
relatively even to eliminate the release of air from beneath the frame.
     Figure 12. Air gap beneath tunnel wall is sealed with soil.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 24 of 3 6
       The test area soil must be free of rocks, twigs, roots, and other objects larger than ~l/2-
inch in any dimension to prevent the rake mechanism from snagging and stalling. The RAFS can
operate on soil with light to medium vegetation. A general rule, the location is acceptable if the
bare soil to vegetation ratio is 1:1 with even coverage. Heavy vegetation, like a lawn, will require
removal of vegetation to achieve the 1:1 ratio (Figure 13). Extremely compacted soil may be
gently loosened with a hand-held pickax if the RAFS rake tines are unable to penetrate the soil.
                    Figure 13. RAFS operating in soil with vegetation.
       The rake is mounted to a shaft, which is  attached to a motorized screw-drive. The
mounting connection  allows for a moderate degree of lateral motion as the rake traverses the
tunnel. The lateral motion of the rake increases the surface area of the soil that is mechanically
disturbed. As the rake traverses  the tunnel, grooves  are created  in  the  soil  and  the rake
mechanism is lowered by its weight to accommodate the soil plane. Typically, the rake reaches a
depth of no less than approximately 1A -inch (Figure 14).

       If precipitation is forecasted, the test locations should be covered with approximately a 6-
ft by 6-ft plastic tarp, which  is held in place with tent  stakes to  maintain the  soil  moisture
content. Measurable particle aerosolization by the  RAFS  does not  occur when  soil  moisture
content is greater than 35%. At this  moisture content, the soil particles  remained agglomerates
because the capillary force induced by the thin water layer between individual particles exceeded
the disturbance force induced by the RAFS.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 25 of 3 6


Figure 14. Proper contact of rake tines with soil resulting in grooves (-^-inch depth).

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 26 of 3 6
4.4   Aerosol Concentration/Sample Volume

       Collection of valid air samples with the RAFS depends on the soil emission rate and
RAFS  filter sample volume (sampling flow rate times sampling time). The sample  collection
time must be sufficiently long to maximize the sample volume collected without overloading the
filter with particles. A high sample volume will reduce the cost of achieving the transmission
electron microscopy (TEM) analytical detection limit (e.g., 0.005 structures/cm3). However, the
filter sample volume must be  balanced against the particle emission rate so the filters are not
overloaded.  Filters  overloaded  with particles  cannot be  analyzed  by direct-transfer TEM
analysis; an indirect transfer method must be used instead.

       Two approaches  are used to determine the paniculate  loading on the filter  sample is
acceptable for analysis of asbestos using direct-transfer TEM.

       Approach 1—first approach involves on-site examination of filter samples using a phase
       contrast microscope (PCM) in accordance  with  NIOSH  Method  7400 to determine
       whether the particulate loading would be  acceptable for subsequent analysis by direct
       transfer TEM. Note: This approach may not be cost reasonable unless the RAFS user is a
       trained optical microscopist.

       Approach 2—second  approach  involves  on-site  visual examination  of  the  filter
       particulate loading with the unaided eye.

       Both approaches use the same basic steps to determine the sample volume that will result
in an acceptable particulate loading. These steps are outlined below:

       1.  Use a hand-held optical particle counter (e.g., MetOne GT-521) to determine the
          particulate  air concentration  (i.e.,  number  particles/cm3) generated  by the RAFS.
          Generally, the highest concentrations are  generated within first 5-minutes of sample
          collection. Hence, the information should be used to determine the minimum sample
          volume for the first of the three samples.

          Table 5 presents guidance to determine a target sampling time (minutes) based on the
          RAFS generated number concentration (number of  particles/cm3) to obtain an
          acceptable  filter  loading  for  asbestos  analysis using  direct  transfer  specimen
          preparation for TEM analysis.  This guidance is based on observing  the number
          concentration profile  over the initial 5-minutes of sampling.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 27 of 3 6
Table 5. Guidance to Determine Target Sampling Time based on RAFS Generated
        Particulate Number Concentration and Sample Flow of 13.5 Lpm.
Aerosol Concentration,
#/cm3
>25
-10
~ C
l~~~> 3
~ 9
/~~~> z*
~ i

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 28 of 3 6
           After achieving the target sampling period, stop the sampling pump and remove the
           filter cassette from location #3. Properly store the filter  cassette for potential
           subsequent TEManalysis in the laboratory.

       6.   Sample collection is now complete.
 Figure 15. Photographs of 25 mm dia. 0.8 um pore MCE filters to assess proper loading of
 particles for direct-transfer TEM analysis, a) unused filter, b) lightly loaded filter, c) ideal,
           properly loaded filter, d) overloaded filter, e) very overloaded filter.
4.5   Loading/Unloading Filter Cassettes

Loading

       Install the site-specific labeled filter cassettes in the  holder located  at the exit of the
RAFS tunnel. The holder supports up to three filter cassettes with the centerline of each cassette
positioned at 1.5-inches above the test surface (Figure 5). The leading edge of the filter cassette
should be placed approximately 2 inches inside the tunnel face. Remove the face cap and outlet
plug from the filter cassette. Attach the vacuum hose to the outlet ferrule of the cassette. Secure
the filter cassette in the mounting  bracket with the screw clamp. Repeat for each cassette.

Unloading

       Loosen the screw clamp and remove the filter cassette. Install face cap  and outlet plug.
Disconnect  the  vacuum hose from the filter cassette.  Exercise sufficient  caution during
disconnecting the filter from the filter cassette not to separate the base of the filter cassette from
the 50 mm extension cowl. It this  occurs, the filter sample should be discarded.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 29 of 3 6
             5   Step-By-Step RAFS Operation
1.    Assemble the RAFS unit.

2.    Set the RAFS operating  settings; see  Section 4.2. Variations  in test matrix surface
     characteristics may directly influence the air flow velocity exit the tunnel. Therefore,
     set the fan speed operating setting on a surface as representative of the test locations.

3.    Clean the RAFS unit; see Section 6.

4.    If necessary, collect an equipment blank and field blanks; see Section 6.

5.    Select/prepare a test location; see Section 4.3.

6.    Place the RAFS on the test location; see Section 4.3.

7.    Remove face cap and outlet plug from  three site-specific labeled 25 mm diameter, 0.8
     |im pore size MCE filter cassettes with  50 mm extension cowls.  Install the filter
     cassettes in the RAFS sampling table.

8.    Position the optical particle counter at the same location as the MCE filter cassettes.

9.    In the following immediate sequential order start the sampling pumps, optical particle
     counter, RAFS fan, and RAFS rake mechanism to collect a test sample.

10.   Sample duration varies with each test location; see Section 4.4.

11.   Record the sampling data and conditions.

12.   Repeatedly check the RAFS system during the sampling period to verify operation of
     the raking mechanism, as well as fan and sampling pumps.

13.   In the following immediate sequential order stop the equipment when the test is being
     stopped or paused (see Section 4.4): sampling pumps, optical  particle  counter, fan, and
     rake mechanism.

14.   Record the sampling data and conditions.

15.   After removing the sample filter cassettes  but prior to moving the RAFS measure the
     air velocity through the tunnel; see  Section 4.2. Prior  to measuring the air velocity,
     carefully redistribute any soil that has accumulated "piled"  at the exit end of the tunnel.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 30 of 36
16.   Clean the RAFS with pre-moistened lint free cloth (or lint free cloth moistened with a
     surface cleaner). Pay particular attention to the inside of the RAFS tunnel,  rake
     mechanism, and filter cassette bracket.

17.   Repeat steps 3 through 16 for collection of the next sample.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 31 of 36
   6     Maintenance, Quality Control, and Calibration
6.1  Cleaning the RAFS

      The RAFS must be thoroughly cleaned both immediately before and after use to remove
all asbestos  and non-asbestos particulate surface contamination. Particular attention should be
paid to the tunnel, raking mechanism, and filter cassette support bracket. The entire RAFS unit
should be thoroughly wiped with pre-moistened lint free cloth (or lint free cloth moistened with a
surface cleaner). Change the cloth frequently to prevent cross-contamination of the surfaces; e.g.,
interior of tunnel vs. exterior.
6.2  Quality Control

RAFS Field Blank

      A field blank is  collected to  check for accidental contamination of samples by the
operator. Follow standard procedures for collection, handling, and processing of the field blank.

RAFS Equipment Blank

      An equipment blank is a quality control sample used to detect asbestos fibers introduced
to  samples  during  the  measurement  process;  i.e.,  it determines  the  cleanliness  and
decontamination effectiveness of the RAFS. Immediately prior to collection of field samples a
RAFS equipment blank sample should be collected to demonstrate the cleanliness of the RAFS
equipment.

      Place the RAFS unit on a clean aluminum (or equivalent non-porous surface) sheet. An
equipment blank (open-face, 25 mm dia., 0.8 jim pore size MCE filter with 50 mm extension
cowl) is collected using exactly the same procedure as used to collect a test sample (Figure 8).
Operate the fan at maximum speed during the sampling period of 5 minutes (minimum). During
the sampling period determine the emission of any particles using a optical particle counter. If
the concentration on the optical particle counter exceeds  100 number of particles/cm3, re-clean
the unit. Replace the MCE filter and begin the test again.

6.3  Equipment  Performance Check and  Calibration

RAFS Unit

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 32 of 36
       The RAFS requires a performance check of its primary operating settings: fan speed and
rake speed. See Section 4.2 "Operating Settings."

Auxiliary Equipment

       A performance check and/or calibration should be performed on each piece of primary
auxiliary equipment in  accordance with  the  respective  operation manual. This includes the
vacuum sampling pumps and optical particle counter, and air flow meter.

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 33 of 36
                       7    Troubleshooting
      Table 6 lists the symptoms, possible causes, and recommended solutions for common
problems that may occur with the RAFS.

                      Table 6. Troubleshooting for RAFS
Symptom
No power to RAFS
No movement of rake
mechanism
Fan does not operate
Vacuum pumps not on
Possible Cause(s)
No power
No power to RAFS
Rake power switch off
GFI tripped
Rake mechanism jammed
No power to RAFS
Fan power switch off
GFI tripped
No power
Corrective Action
Check generator or 1 10 VAC
outlet
Check power source; switch on
Switch on
Reset GFI switch
Inspect mechanism for
obstruction. Remove obstruction
Check power source; switch on
Switch on
Reset GFI switch
Check power cords
Check power switch
Check power source

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 34 of 36
                           8     References
1.  Kominsky, J., Thornburg, J, Shaul, G., Barrett, W., and Hall, F. Development of the
   Releasable Asbestos Field Sampler. J. Air Waste Management Assoc. in press.

2.  Hinds, W. Aerosol Technology. 2nd Ed., John Wiley & Sons. 1999.

3.  International Organization for Standardization. ISO Method 10312:1995. Ambient Air-
   Determination of Asbestos Fibres—Direct Transfer Transmission Electron Microscopy
   Method (1995).

-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 35 of 36
                  Appendix A:  Example Forms
Example RAFS Filter Sample Label
      RAFS-A#-L#-X-#
   RAFS-O = Outdoor RAFS
   A# = Area Number (e.g., A1)
   L# = Location # in that Area (e.g., L1)
   X = Type of sample (S = Sample, B = Field blank, E = Equipment
       blank, D = Duplicate)
   # = Location in RAFS sample manifold (1, 2, or 3)	
         Sample ID
     AIR SAMPLE CUSTODY FORM

RELEASABLE ASBESTOS FIELD SAMPLER (RAFS)

      Tampa, FL: GA008 and GA0108
    Sample Type  Volume (L)  Shipper ID
Date   Receipt ID  Date
GAO08-I-RAFS-A1 -L1 -01
GAO08-I-RAFS-A1 -L1 -02
GA008-I-RAFS-A1-L3-01
GAO08-I-RAFS-A1 -L3-02
Sample
Field Blank
Sample
Duplicate
100
0
50
50
AD
AD
AD
AD
9/14/2009
9/14/2009
9/14/2009
9/14/2009









-------
QLOG#S-14397-OP-1-0
REVISION: 0
DATE: 9/28/2011
Page 36 of 36
RAFS Datasheet Date:
AIR SAMPLING AND INSTRUMENT DATA COLLECTION FORM Tech:
RELEASABLE ASBESTOS FIELD SAMPLER (RAFS) GPS: N
Sampling Condition Parameters
Area:
Weather Phenomena
Location Description
W
Altitude: feef

Location: RH: % Temp: °F WindVel: miles/hr Bar Press: inches of Hg
(e.g. rain, wind gusts): Direction: DewPt: °F
(e.g. vegetation, soil consistency): Test Qualifier:
Sample
ID






MetOneGT 521 Filename


RAFS
Unit N2






Start


Pump N2






Stop


Flow Rate Agitator Data Time RJI
Start Stop Start Stop Cycle Depth Comments Filter
(Urn) (Urn) Count Count Time(s) (inches) P ID






Flow
_ . ... . Comments Vane Anemometer
Rate (L/m)
feet/minute

Soil Moisture Readings Soil Sample Tracking
Reading 1

Reading 2

Reading 3

Reading 4

Reading 5

Split # Sample ID Comments
1
Additional Sample Period Comments
3
4
5

-------