SEPA
United States
Environmental Protection
Agency
 www.epa.gov
  Method 614: The Determination of
  Organophosphorus Pesticides in
  Municipal and Industrial
  Waste water

-------
               Method 614
         The Determination of
Organophosphorus Pesticides in
      Municipal and Industrial
                  Wastewater

-------
                                 Method 614
 The Determination of Organophosphorus Pesticides in Municipal and
                             Industrial Wastewater


1.     SCOPE AND APPLICATION

1.1    This method covers the determination of certain organophosphorus pesticides.  The
      following parameters can be determined by this method:

          Parameter                    STORET No.            CAS No.
          Azinphos methyl                 39580                86-50-0
          Demeton                        39560              8065-48-3
          Diazinon                        39570               333-41-5
          Disulfoton                       39010               298-04-4
          Ethion                            -                 563-12-2
          Malathion                       39530               121-75-5
          Parathion ethyl                  39540                56-38-2
          Parathion methyl                 39600               298-00-0

1.2    This is a gas chromatographic  (GC) method  applicable  to  the determination of the
      compounds listed  above in industrial and  municipal discharges as provided under
      40 CFR 136.1. Any modification of this method beyond those expressly permitted shall
      be considered a major modification subject to application and approval of alternative test
      procedures under 40 CFR 136.4 and 136.5.

1.3    The method detection limit (MDL, defined in Section 15) for several parameters are listed
      in Table  1.  The MDL for a specific wastewater may differ from those listed, depending
      upon the nature of interferences in the sample  matrix.

1.4    The sample extraction and concentration steps in this method are essentially the same as
      in Method 617.  Thus, a single sample may be  extracted to measure the parameters
      included in the scope of both of these  methods.   When  cleanup is required, the
      concentration levels must be high enough  to permit selecting aliquots, as necessary, in
      order to apply appropriate cleanup procedures. Under gas chromatography, the analyst
      is allowed  the latitude  to select  chromatographic conditions  appropriate  for the
      simultaneous measurement of combinations of these parameters (see Section 12).

1.5    This method is restricted to use by or under the supervision of analysts experienced in
      the use of gas chromatography and in the interpretation of gas chromatograms.  Each
      analyst must demonstrate the ability to generate acceptable results  with this method
      using the procedure described in Section 8.2.

1.6    When this method is used to analyze unfamiliar samples for any or all of the compounds
      above,  compound identifications  should be  supported  by  at least  one additional
      qualitative  technique.   This method describes analytical  conditions for a second gas
      chromatographic column that can be used to confirm measurements made with the

-------
Method 614
       primary column. Section 14 provides gas chromatograph/mass spectrometer (GC/MS)
       criteria appropriate for the qualitative confirmation of compound identifications.

2.     SUMMARY OF METHOD

2.1     A measured volume of sample,  approximately 1  L, is  extracted with 15% methylene
       chloride in hexane using a separatory funnel.  The extract is dried and concentrated to
       a volume of 10 mL or less. Gas chromatographic conditions are described which permit
       the separation and measurement of the compounds in the extract by flame photometric
       or thermionic bead gas chromatography.

2.2     Method 614 represents  an editorial revision  of a previously promulgated U.S. EPA
       method for organophosphorus pesticides.1  While complete method validation data is not
       presented herein, the method has been in widespread use since its promulgation, and
       represents the state of the art for the analysis of such materials.

2.3     This  method provides  selected  cleanup procedures  to  aid  in the elimination  of
       interferences which may be encountered.

3.     INTERFERENCES

3.1     Method interferences may be caused by contaminants in solvents, reagents, glassware,
       and other sample-processing apparatus that lead to  discrete artifacts or elevated baselines
       in gas chromatograms. All reagents and apparatus must be routinely demonstrated to
       be free from interferences under the conditions of the analysis by running laboratory
       reagent blanks as described in Section 8.5.

       3.1.1    Glassware must be scrupulously cleaned.2 Clean all glassware as soon as possible
              after use by thoroughly rinsing with  the  last solvent used in it.   Follow  by
              washing with hot water and detergent and thorough rinsing with tap and reagent
              water.  Drain dry, and heat in an  oven or  muffle furnace at 400°C for 15 to 30
              minutes. Do not heat volumetric ware.  Thermally stable materials, such as PCBs,
              may not be eliminated by this treatment.  Thorough rinsing with acetone and
              pesticide-quality  hexane  may be substituted for  the heating.  After drying and
              cooling,  seal  and store glassware  in  a  clean  environment to prevent any
              accumulation of  dust or  other contaminants.  Store inverted or capped with
              aluminum foil.

       3.1.2    The use of  high-purity reagents and solvents helps to  minimize interference
              problems.  Purification of solvents by distillation in all-glass systems may be
              required.

3.2     Matrix interferences may be  caused by  contaminants  that  are  coextracted from the
       sample. The extent of matrix interferences will vary considerably from source to source,
       depending upon the nature  and  diversity  of the industrial complex  or municipality
       sampled.  The cleanup procedure in Section 11  can be used to overcome many of these
       interferences,  but unique samples may require additional cleanup approaches to achieve
       the MDL listed in Table  1.

-------
                                                                                Method 614
4.     SAFETY

4.1     The toxicity or carcinogenicity of each reagent used in this method has not been precisely
       defined; however, each chemical compound must be treated as a potential health hazard.
       From this viewpoint, exposure to these chemicals must be reduced to the lowest possible
       level by whatever means available.  The laboratory is responsible for maintaining a
       current awareness file  of OSHA regulations regarding the safe handling of the chemicals
       specified in this method. A reference file of material data handling sheets should also
       be made available  to all personnel involved  in the  chemical  analysis.   Additional
       references to laboratory  safety are  available and have  been identified35  for  the
       information of the analyst.

5.     APPARATUS AND MATERIALS

5.1     Sampling equipment,  for discrete or composite sampling.

       5.1.1  Grab-sample bottle:  Amber borosilicate or  flint glass, 1-L or 1-quart volume,
             fitted with screw-caps lined with TFE-fluorocarbon.  Aluminum foil may  be
             substituted for TFE if the sample is not corrosive.   If amber  bottles are  not
             available, protect samples from light.   The container and  cap liner must  be
             washed,  rinsed with acetone or methylene  chloride, and dried before  use to
             minimize contamination.

       5.1.2  Automatic sampler (optional):  Must incorporate glass sample containers for the
             collection of a minimum of 250 mL. Sample containers must be kept refrigerated
             at 4°C and protected from light during compositing.  If the sampler  uses a
             peristaltic pump, a minimum length of compressible silicone  rubber tubing may
             be used.  Before use, however, the compressible tubing must be thoroughly rinsed
             with methanol, followed by repeated rinsings  with reagent water  to minimize the
             potential for  contamination of the sample. An integrating  flow meter is required
             to collect flow-proportional composites.

5.2     Glassware.   (All  specifications are  suggested.   Catalog numbers are  included for
       illustration only.)

       5.2.1  Separatory funnel:   125-mL, 1000-mL,  and 2000-mL, with TFE-fluorocarbon
             stopcock, ground-glass or TFE stopper.

       5.2.2  Drying column:   Chromatographic column  400 mm  long by 19 mm ID with
             coarse-fritted disc.

       5.2.3  Chromatographic column:  400 mm long by 19 mm ID with coarse-fritted disc at
             bottom and TFE-fluorocarbon stopcock  (Kontes  K-420540-0224 or equivalent).

       5.2.4  Concentrator tube, Kuderna-Danish: 10-mL, graduated (Kontes K-570050-1025 or
             equivalent).  Calibration must be checked at  the volumes employed in the test.
             Ground-glass stopper is used to prevent evaporation of extracts.

-------
Method 614
       5.2.5  Evaporative  flask,  Kuderna-Danish:    500-mL   (Kontes  K-570001-0500  or
             equivalent).  Attach to concentrator tube with springs.

       5.2.6  Snyder column, Kuderna-Danish:  Three-ball macro  (Kontes K-503000-0121 or
             equivalent).

       5.2.7  Snyder column, Kuderna-Danish:  Two-ball micro (Kontes K-569001-0219 or
             equivalent).

       5.2.8  Pipette, disposable:  140 mm long by  5 mm ID.

       5.2.9  Vials:  Amber glass, 10- to 15-mL capacity with TFE-fluorocarbon-lined screw-cap.

5.3     Boiling chips:  Approximately  10/40 mesh.  Heat at  400°C for 30 minutes or perform a
       Soxhlet extraction with methylene chloride.

5.4     Water bath:  Heated, with concentric ring cover, capable of temperature control (±2°C).
       The bath should be  used  in a hood.

5.5     Balance:  Analytical, capable of accurately weighing  to the nearest 0.0001 g.

5.6     Gas chromatograph: Analytical system complete with gas chromatograph suitable for
       on-column injection and all required accessories including syringes, analytical columns,
       gases, detector, and  strip-chart  recorder.  A data system is recommended for measuring
       peak areas.

       5.6.1  Column 1: 180 cm long by 4 mm ID glass, packed  with 3% OV-1 on Gas Chrom
             Q (100/120 mesh)  or equivalent.  This column was used to develop the method
             performance statements in  Section 15.   Alternative columns may be used in
             accordance with the provisions described in Section 12.1.

       5.6.2  Column 2: 180 cm long by 4 mm ID glass, packed with 1.5% OV-17/1.95% QF-1
             on Gas Chrom Q (100/120 mesh) or equivalent.

       5.6.3  Detector: Phosphorus-specific; flame photometric  detector (FPD,  with 526 nm
             filter) or thermionic bead detector in  the nitrogen  mode.  These detectors have
             proven effective in the  analysis of wastewaters for the parameters listed in the
             scope.  The  FPD was used to  develop  the method performance statements in
             Section 15. Alternative detectors,  including a  mass  spectrometer, may be used in
             accordance with the provisions described in Section 12.1.

6.     REAGENTS

6.1     Reagent water:  Reagent water is  defined as  a water  in which an interferent  is not
       observed at the method detection limit of each parameter of interest.

6.2     Acetone, hexane, isooctane, methylene chloride: Pesticide-quality or equivalent.

-------
                                                                                 Method 614
6.3     Ethyl ether: Nanograde, redistilled in glass if necessary.  Must be free of peroxides as
       indicated  by  EM  Quant  test  strips (available  from  Scientific  Products Co.,  Cat.
       No. PI 126-8, and other suppliers). Procedures recommended for removal of peroxides
       are provided with the test strips. After cleanup, 20 mL of ethyl alcohol preservative must
       be added to each liter of ether.

6.4     Acetonitrile, hexane-saturated: Mix pesticide-quality acetonitrile with an excess of hexane
       until equilibrium is established.

6.5     Sodium sulfate:  ACS, granular, anhydrous.  Condition by heating in  a shallow tray at
       400°C  for a minimum of 4 hours to  remove phthalates and other interfering organic
       substances.  Alternatively, heat 16 hours at  450-500°C  in a shallow tray or perform a
       Soxhlet extraction with methylene chloride for 48 hours.

6.6     Sodium chloride solution, saturated: Prepare saturated solution of NaCl in reagent water
       and extract with hexane to remove impurities.

6.7     Alumina:  Woelm, neutral; deactivate  by  pipetting 1 mL of distilled water into a 125-mL
       ground-glass stoppered Erlenmeyer flask.  Rotate flask  to distribute water over surface
       of glass. Immediately add 19.0 g fresh alumina through small powder funnel.  Shake
       flask containing mixture for 2 hours on a mechanical shaker.

6.8     Florisil: PR grade (60/100 mesh).  Purchase activated at 675°C and store in dark in glass
       container with ground-glass stopper  or foil-lined screw-cap.  Before use, activate each
       batch at least 16 hours at  130°C in a foil-covered glass container.

6.9     Stock standard solutions (1.00 ug/uL): Stock standard solutions may be prepared from
       pure standard materials or purchased as certified solutions.

       6.9.1   Prepare stock standard solutions by accurately weighing approximately 0.0100 g
              of pure material.  Dissolve the material in pesticide-quality isooctane or acetone
              and dilute to volume in a 10-mL volumetric flask.  Larger volumes may be used
              at the convenience of the  analyst. If compound purity is certified at 96% or
              greater, the weight may be used without correction to calculate  the concentration
              of the stock standard. Commercially prepared stock standards may be used at
              any concentration if they are certified  by the manufacturer or by an independent
              source.

       6.9.2   Transfer  the stock  standard solutions into TFE-fluorocarbon-sealed screw-cap
              vials.   Store at 4°C and protect  from light. Frequently check stock standard
              solutions  for  signs of degradation  or evaporation,  especially  just prior to
              preparing calibration standards from  them.

       6.9.3   Stock standard solutions must be replaced after 6 months, or sooner if comparison
              with check standards indicates a problem.

-------
Method 614
7.     CALIBRATION

7.1     Establish gas chromatographic operating parameters equivalent to those indicated in
       Table 1.  The gas chromatographic system may be calibrated using either the external
       standard technique (Section 7.2) or the internal standard technique (Section 7.3).

7.2     External standard calibration procedure.

       7.2.1   For each  parameter of interest, prepare calibration standards at a minimum of
              three concentration levels by adding accurately measured volumes of one or more
              stock standards to a volumetric flask and diluting to volume with isooctane or
              other suitable solvent.  One of the external standards should be representative of
              a  concentration  near, but above, the  method detection  limit.   The  other
              concentrations should correspond to the range of concentrations expected in the
              sample concentrates or should define the working range of the  detector.

       7.2.2   Using injections of 1 to 5 uL of each calibration standard, tabulate peak height or
              area responses against the  mass injected.  The results can be used to prepare a
              calibration curve for each parameter.  Alternatively, the ratio of the  response to
              the  mass injected,  defined  as the  calibration factor (CF), may be calculated for
              each parameter at each standard concentration. If the relative standard deviation
              of the calibration factor is  less than  10% over the working range, the average
              calibration factor can be used in place of a calibration curve.

       7.2.3   The working calibration curve or calibration factor must be verified on  each
              working shift by the measurement of one or more calibration standards. If the
              response  for any parameter varies from the predicted  response  by more  than
              ±10%, the test must be repeated using a fresh calibration  standard.  Alternatively,
              a new calibration curve or calibration factor must be prepared for that parameter.

7.3     Internal standard calibration procedure:  To use this approach, the analyst must select
       one or more internal standards similar  in  analytical behavior to the compounds of
       interest.  The analyst must further demonstrate that the measurement of the internal
       standard is not affected by method or matrix interferences.  Due to these  limitations, no
       internal standard applicable to all  samples can be suggested.

       7.3.1   Prepare calibration standards at a minimum of three concentration  levels for each
              parameter of interest by adding volumes of one or more stock standards to a
              volumetric flask. To each calibration standard, add a known constant amount of
              one or more internal standards, and dilute to volume  with isooctane or  other
              suitable solvent.  One of the standards should be representative of a concentration
              near, but above, the method detection limit.  The other concentrations should
              correspond to the range of concentrations expected in the sample concentrates, or
              should define the working  range of the detector.

       7.3.2   Using injections of 1 to  5 uL of each calibration standard,  tabulate the peak height
              or area responses against  the concentration for each  compound and internal
              standard.  Calculate response factors (RF) for each compound as follows:

-------
                                                                                Method 614
                                       Equation 1


                                     RF  = —	
              where
              As = Response for the parameter to be measured
              Als = Response for the internal standard
              Cls = Concentration of the internal standard, in ug/L
              Cs = Concentration of the parameter to be measured, in ug/L
              If the RF value over the working range is constant, less than  10%  relative
              standard deviation, the RF can be assumed to be invariant and the average RF
              may be  used for calculations.  Alternatively, the results may be used to plot a
              calibration curve of response ratios, AS/A is against RF.

       7.3.3   The working calibration curve or RF must be verified on each working shift by
              the  measurement of one or more calibration standards. If the response for any
              parameter varies from the predicted response by more  than ±10%,  the test must
              be repeated using a fresh calibration standard.  Alternatively, a new calibration
              curve must be prepared for that compound.

7.4     The cleanup procedure in Section  11 utilizes Florisil  chromatography.   Florisil from
       different batches or sources may vary in adsorptive capacity.  To standardize the amount
       of Florisil which is used, the use of the  lauric acid value is suggested.  This procedure6
       determines the  adsorption from hexane solution of lauric acid, in milligrams, per gram
       of Florisil.  The amount of Florisil to be used for each column  is calculated by dividing
       this factor into  110 and multiplying by 20 g.

7.5     Before using any cleanup procedure, the analyst must process a series of calibration
       standards  through  the procedure  to validate  elution patterns and the  absence  of
       interference from the reagents.

8.     QUALITY CONTROL

8.1     Each  laboratory using this method  is required to operate a formal quality control
       program. The minimum requirements  of this program consist of an initial demonstration
       of laboratory capability and the analysis of spiked samples as a continuing check  on
       performance. The laboratory is required to maintain performance records to define the
       quality of data  that is generated.

       8.1.1   Before performing any analyses, the analyst must  demonstrate  the ability to
              generate acceptable accuracy  and precision with this  method.  This ability is
              established as described in Section 8.2.

-------
Method 614
       8.1.2  In recognition of the rapid advances occurring in chromatography, the analyst is
             permitted  certain options to improve the separations or lower  the  cost of
             measurements. Each time such modifications to the method are made, the analyst
             is required to repeat the procedure in Section  8.2.

       8.1.3  The laboratory must spike and analyze a minimum of 10%  of all samples to
             monitor continuing laboratory performance.   This procedure is described in
             Section 8.4.

8.2     To establish the ability to generate acceptable accuracy and precision, the analyst must
       perform the following operations.

       8.2.1  Select a representative spike concentration for each compound to be measured.
             Using  stock  standards, prepare a quality  control check sample concentrate in
             acetone, 1000 times more concentrated than the selected concentrations.

       8.2.2  Using  a pipette, add 1.00 mL of the check  sample concentrate to each  of a
             minimum  of four  1000-mL  aliquots  of reagent  water.    A representative
             wastewater may be used in place of the reagent water, but one or more additional
             aliquots must be analyzed to  determine background levels, and the spike level
             must exceed twice the  background level for the  test to be valid.  Analyze the
             aliquots according to the method beginning in Section 10.

       8.2.3  Calculate the average percent recovery (R), and  the standard deviation of the
             percent recovery (s), for the results.  Wastewater background corrections must be
             made before R and calculations are performed.

       8.2.4  Table 2 provides single-operator recovery and precision for diazinon, parathion
             methyl, and  parathion  ethyl.  Similar results  should be expected from  reagent
             water for all organophosphorus compounds listed in this method.  Compare these
             results to the values calculated in Section  8.2.3. If the data are not comparable,
             review  potential problem areas and repeat the test.

8.3     The analyst must calculate method performance criteria and define the performance of
       the laboratory  for each spike concentration and parameter being measured.

       8.3.1  Calculate upper and lower control limits for method performance as follows:

                    Upper Control Limit (UCL)  = R + 3s
                    Lower Control Limit (LCL)  = R - 3s

             where R and S are calculated as in Section 8.2.3.  The UCL and LCL can be used
             to construct control charts7 that are useful in observing trends in performance.

       8.3.2  The laboratory  must develop and maintain  separate  accuracy  statements of
             laboratory performance for wastewater samples. An accuracy statement for the
             method is defined as R  ± s. The accuracy statement should be developed by the
             analysis of four aliquots of wastewater as described in Section  8.2.2,  followed by
             the calculation of R and s. Alternatively, the  analyst may use four wastewater

-------
                                                                               Method 614
             data points gathered through the requirement for continuing quality control in
             Section 8.4. The accuracy statements should be updated regularly.7

8.4     The laboratory is required to collect in duplicate a portion of their samples to monitor
       spike recoveries.  The frequency of spiked sample analysis must be at least 10% of all
       samples or one spiked sample  per  month, whichever is  greater. One aliquot of the
       sample must be spiked and analyzed as described in Section 8.2. If the recovery for a
       particular parameter does not fall within the control limits for method performance, the
       results reported for that parameter in all samples processed as part of the same set must
       be qualified as described in Section 13.3. The laboratory should monitor the frequency
       of data so  qualified to ensure that it remains at or below 5%.

8.5     Before processing any samples, the analyst must demonstrate through the  analysis of a
       1-L  aliquot of reagent water that all glassware and reagent interferences  are under
       control.  Each time  a set of samples is extracted  or there is  a  change in reagents, a
       laboratory reagent  blank must  be processed as  a  safeguard against  laboratory
       contamination.

8.6     It is recommended that the laboratory adopt additional quality assurance  practices for
       use with this method. The specific practices that are  most productive depend upon the
       needs of the laboratory and the nature of the samples.  Field duplicates may be analyzed
       to monitor the precision of the sampling technique.  When doubt exists over  the
       identification  of a peak on the chromatogram, confirmatory techniques  such as  gas
       chromatography  with a  dissimilar column,  specific  element detector,  or  mass
       spectrometer must be used.  Whenever possible, the laboratory should perform analysis
       of quality control materials and  participate in relevant performance evaluation studies.

9.     SAMPLE COLLECTION, PRESERVATION, AND HANDLING

9.1     Grab samples must be collected in glass containers.  Conventional sampling practices8
       should be followed;  however, the bottle must not  be  prerinsed with sample before
       collection.  Composite samples  should  be  collected in refrigerated glass containers in
       accordance with the requirements of the  program. Automatic sampling equipment must
       be as free as possible of plastic and other potential sources of contamination.

9.2     The samples  must be iced or  refrigerated at 4°C from the  time of collection until
       extraction.

9.3     All samples must be extracted within 7 days and completely analyzed within 40 days of
       extraction.

10.    SAMPLE EXTRACTION

10.1    Mark the water meniscus on the side of the sample bottle for later  determination of
       sample volume.  Pour the entire sample into a 2-L separatory funnel.

10.2    Add 60 mL of 15% (v/v) methylene chloride in hexane to the sample bottle, seal, and
       shake 30 seconds to rinse the inner walls. Transfer the solvent to  the separatory funnel
       and extract the sample by shaking the  funnel  for 2  minutes with periodic venting to

-------
Method 614
       release excess pressure. Allow the organic layer to separate from the water phase for a
       minimum of 10 minutes. If the emulsion interface between layers is more than one third
       the volume of the solvent layer, the analyst must employ mechanical techniques to
       complete the phase separation. The optimum technique depends upon the sample, but
       may  include stirring, filtration of the emulsion  through glass wool, centrifugation, or
       other physical methods.  Drain the aqueous phase into a 1000-mL Erlenmeyer flask and
       collect  the extract in a 250-mL Erlenmeyer  flask.  Return the  aqueous phase to the
       separatory funnel.

10.3    Add  a second 60-mL volume of 15% methylene chloride in hexane to the sample bottle
       and repeat the extraction procedure a second time, combining the extracts in the 250-mL
       Erlenmeyer flask.  Perform a third extraction in the same manner.

10.4    Assemble a Kuderna-Danish  (K-D) concentrator by attaching a 10-mL concentrator tube
       to a 500-mL evaporative flask.  Other concentration devices or techniques may be used
       in place of the K-D if the requirements of Section 8.2 are met.

10.5    Pour the combined  extract  through  a drying  column  containing about 10 cm of
       anhydrous sodium sulfate, and collect the extract in  the K-D  concentrator.  Rinse the
       Erlenmeyer flask and column with 20 to 30 mL  of hexane to complete  the quantitative
       transfer.

10.6    Add  one or two clean boiling chips to the  evaporative flask and attach a three-ball
       Snyder column. Prewet the Snyder column by adding about 1 mL methylene chloride
       to the  top.   Place  the  K-D apparatus  on a hot water bath,  80 to  85°C, so that the
       concentrator tube is partially immersed in the hot water, and the entire lower rounded
       surface of the flask is bathed with hot  vapor.  Adjust the  vertical  position of the
       apparatus and the water temperature as required to complete the concentration in 15 to
       20 minutes.  At the proper rate of distillation, the balls of the column will actively chatter
       but the chambers will not flood with condensed solvent. When the apparent volume of
       liquid reaches 1 mL, remove the K-D apparatus and allow it to drain and cool for at least
       10 minutes.

10.7    Remove the Snyder column and rinse the flask and its lower joint into the concentrator
       tube  with 1 to 2 mL of hexane and adjust the volume to 10  mL.  A  5-mL syringe is
       recommended for this operation. Stopper the concentrator tube and store refrigerated
       if further processing will not be performed immediately.  If the  extracts will be stored
       longer than two days, they should be transferred to PTFE-sealed screw-cap bottles.  If the
       sample extract requires no further cleanup, proceed with gas chromatographic analysis.
       If the sample requires cleanup, proceed to Section 11.

10.8    Determine the original sample volume by refilling the sample bottle to the mark and
       transferring the water to a 1000-mL graduated cylinder.  Record the sample volume to
       the nearest 5 mL.

11.    CLEANUP AND SEPARATION

11.1    Cleanup procedures may not be necessary for a relatively clean sample matrix.  The
       cleanup procedure recommended in this  method  has  been used for the  analysis of

-------
                                                                                Method 614
       various industrial and municipal effluents.  If particular circumstances demand the use
       of an alternative cleanup procedure, the analyst must determine the elution profile and
       demonstrate that the recovery of each compound of interest for the cleanup procedure
       is no less than 85%.

11.2    Acetonitrile partition:  The following acetonitrile partitioning procedure may be used to
       isolate fats and oils from the sample  extracts.  The applicability of this procedure to
       organophosphorus pesticides is indicated in Table 3.

       11.2.1 Quantitatively transfer the previously concentrated extract to a 125-mL separatory
             funnel with enough hexane to bring the final volume to 15 mL.  Extract the
             sample 4 times by shaking vigorously for 1 minute with  30-mL  portions of
             hexane-saturated acetonitrile.

       11.2.2 Combine and transfer the acetonitrile phases to a 1-L separatory funnel and add
             650 mL of reagent water and 40 mL of saturated sodium chloride solution.  Mix
             thoroughly  for 30 to 45 seconds. Extract with two 100-mL portions of hexane by
             vigorously shaking for 15 seconds.

       11.2.3 Combine the hexane extracts in a 1 L separatory funnel and wash with  two
             100 mL portions of reagent water.  Discard the water layer and pour the hexane
             layer through a drying column containing 7-10 cm of anhydrous sodium sulfate
             into a 500 mL K-D flask equipped with a 10-mL concentrator tube.  Rinse the
             separatory funnel and column with three 10-mL portions of hexane.

       11.2.4 Concentrate the extracts to 6 tolO  mL in the K-D as directed in Section 10.6.
             Adjust the extract volume to 10 mL with hexane.

       11.2.5 Analyze by gas chromatography unless a need for further cleanup is indicated.

11.3    Florisil column cleanup: The following Florisil column cleanup procedure has been
       demonstrated to be applicable to the seven organophosphorus pesticides listed in Table 3.
       It should also be applicable to the cleanup of extracts for ethion.

       11.3.1 Add a  weight  of Florisil  (nominally 20  g)  predetermined  by calibration
             (Sections 7.4 and  7.5) to a chromatographic column.  Settle the Florisil by tapping
             the column. Add anhydrous sodium sulfate to the top of the Florisil to form a
             layer 1 to 2  cm deep.  Add 60 mL of hexane to wet and rinse the sodium sulfate
             and Florisil. Just prior to exposure of the sodium sulfate to air, stop the elution
             of the hexane by closing the stopcock on the chromatography column.  Discard
             the eluate.

       11.3.2 Adjust the sample extract volume to 10 mL with hexane and transfer it from the
             K-D concentrator tube to the Florisil column. Rinse the tube twice with 1 to 2 mL
             hexane, adding each rinse to the column.

       11.3.3 Place a 500-mL K-D flask and clean concentrator tube under the chromatography
             column.  Drain the column into the flask until the sodium sulfate layer is nearly
             exposed.  Elute the column with 200  mL  of 6% (v/v) ethyl ether in hexane

-------
Method 614
              (Fraction 1) using a drip rate of about 5 mL/min. Remove the K-D flask and set
              aside for later concentration.  Elute the column again, using 200 mL of 15% (v/v)
              ethyl ether in hexane (Fraction 2) into a second K-D flask.  Perform a third elution
              using 200 mL of 50% (v/v) ethyl ether in hexane (Fraction 3) and a final elution
              with 200  mL of 100%  ethyl ether  (Fraction 4)  into separate K-D flasks.  The
              elution patterns for seven of the pesticides are shown in Table 3.

       11.3.4  Concentrate the eluates by standard K-D techniques  (Section 10.6), using the
              water bath at about 85°C (75°C for Fraction 4). Adjust final volume to  10 mL
              with hexane.  Analyze by gas chromatography.

11.4    Removal of sulfur:9  Elemental sulfur  will elute  in Fraction  1  of the Florisil cleanup
       procedure.   If a  large  amount of sulfur is present in  the extract, it may elute  in all
       fractions. If so, each fraction must be further treated to remove the sulfur.

       11.4.1  Add one  or two  boiling chips to  the 10-mL  hexane solution  contained in  a
              concentrator tube. Attach a micro-Snyder column and  concentrate the extract to
              about 0.2 mL in a hot water bath at 85°C.  Remove the  micro K-D from the bath,
              cool, and adjust the volume to 0.5 mL with hexane.

       11.4.2  Plug a disposable pipette with a small  quantity of glass  wool. Add enough
              alumina to produce a 3-cm column after settling. Top the alumina with a 0.5-cm
              layer of anhydrous sodium sulfate.

       11.4.3  Quantitatively transfer the concentrated extract to the alumina microcolumn using
              a 100- uL  syringe.  Rinse the ampule with 200 uL of hexane and  add to the
              microcolumn.

       11.4.4  Elute the microcolumn  with 3 mL of hexane and discard the eluate.

       11.4.5  Elute the column with 5 mL of 10% hexane in methylene chloride, and collect the
              eluate in a 10-mL concentrator tube.  Adjust final volume to 10 mL with hexane.
              Analyze by gas chromatography.

12.    GAS CHROMATOGRAPHY

12.1    Table 1 summarizes the recommended operating conditions for the gas chromatograph.
       Included in this table are estimated retention-times and method detection limits that can
       be achieved by this method.  Other packed columns,  chromatographic conditions, or
       detectors may be  used if the requirements  of  Section 8.2  are met.   Capillary
       (open-tubular) columns may also be used if the relative standard deviations of responses
       for replicate injections are demonstrated to  be  less than 6%  and the requirements of
       Section 8.2 are met.

12.2    Calibrate the system daily as described in Section 7.

12.3    If the internal standard approach is being used,  add  the internal standard to sample
       extracts immediately before injection into the instrument.  Mix thoroughly.

-------
                                                                               Method 614
12A    Inject  1 to 5 |_iL of the sample extract using the solvent-flush technique.10 Record the
       volume injected to the nearest 0.05 uL, and the resulting peak size in area or peak height
       units.  An automated system that consistently injects a constant volume of extract may
       also be used.

12.5    The width of the retention-time window used to make identifications should be based
       upon measurements of actual retention-time variations of standards over the course of
       a day.  Three times the standard deviation of a retention time can be used to calculate
       a suggested window size for  a  compound.  However,  the experience of the analyst
       should weigh heavily in the interpretation of chromatograms.

12.6    If the response for the peak exceeds the working range of the system, dilute the extract
       and reanalyze.

12.7    If the measurement of the peak response is prevented  by the presence of interferences,
       further cleanup is required.

13.    CALCULATIONS

13.1    Determine the  concentration of individual compounds in the sample.

       13.1.1  If the external standard calibration procedure  is  used, calculate the amount of
             material injected from the  peak response using the calibration curve or calibration
             factor in Section 7.2.2.  The concentration in the sample can be  calculated as
             follows:
                                       Equation 2


                            Concentration, \ig/L  -
             where
             A  = Amount of material injected, in ng
             V,  = Volume of extract injected, in uL
             Vt  = Volume of total extract, in uL
             V,  = Volume of water extracted, in ml
       13.1.2 If  the  internal  standard  calibration  procedure  was  used,  calculate  the
             concentration in the sample using the response factor (RF)  determined in Section
             7.3.2 as follows:
                                       Equation 3
                                                   (V) (V)

-------
Method 614
                         Concentration, \ig/L -       s   s
                                                (Ais) (RF)
              where
              As  = Response for parameter to be measured
              Als = Response for the internal standard
              Is  = Amount of internal standard added to each extract, in ug
              V0  = Volume of water extracted, in L
13.2    Report results in micrograms per liter without correction for recovery data.  When
       duplicate and spiked samples are analyzed, report all data obtained with the sample
       results.

13.3    For samples processed as part of a set where the laboratory spiked sample recovery falls
       outside of the control limits in Section 8.3, data for  the affected parameters  must be
       labeled as suspect.

14.    GC/MS  CONFIRMATION

14.1    It  is recommended  that  GC/MS techniques  be  judiciously employed  to  support
       qualitative compound identifications made with this  method.  The mass spectrometer
       should be capable of scanning the mass range from 35 amu to a mass 50 amu above the
       molecular weight of the compound.  The instrument must be capable of scanning the
       mass range  at a rate to produce at least 5 scans per peak but not to exceed 7 seconds per
       scan utilizing a 70 V (nominal) electron energy in the  electron impact ionization mode.
       A GC-to-MS interface constructed of all glass or glass-lined materials  is recommended.
       A  computer system should be interfaced to the mass  spectrometer that  allows the
       continuous  acquisition and storage on machine-readable media  of all mass spectra
       obtained  throughout the duration of the chromatographic program.

14.2    Gas chromatographic columns and conditions should be selected for optimum separation
       and performance.  The conditions selected must be compatible with standard GC/MS
       operating practices. Chromatographic tailing factors of less than 5.0 must be achieved.11

14.3    At the beginning  of each day that confirmatory analyses are to be performed,  the GC/MS
       system must be checked  to see that  all  decafluorotriphenyl phosphine  (DFTPP)
       performance criteria are achieved.12

14.4    To confirm  an identification of a compound, the background-corrected mass spectrum
       of the compound must be obtained from the sample extract and compared with a mass
       spectrum from a  stock or calibration standard analyzed under the same chromatographic
       conditions.  It is recommended that at least 25 ng of material be injected into the  GC/MS.
       The criteria below must be met for qualitative confirmation.

-------
                                                                              Method 614
       14.4.1  All ions that are present above 10% relative abundance in the mass spectrum of
             the standard must be present in the mass spectrum of the sample with agreement
             to ±10%.  For example, if the relative abundance of an ion is 30% in the mass
             spectrum of the standard, the allowable limits for the relative abundance of that
             ion in the mass spectrum for the sample would be 20 to 40%.

       14.4.2  The retention-time of the compound in the sample must be within 6 seconds of
             the same compound in the standard solution.

       14.4.3  Compounds that have  very similar mass spectra can be explicitly identified by
             GC/MS only on the basis of retention-time data.

14.5    Where available, chemical ionization mass spectra  may be  employed to  aid in  the
       qualitative identification process.

14.6    Should these MS procedures fail to provide satisfactory results, additional steps may be
       taken before reanalysis. These may include the use of alternate packed or capillary GC
       columns or additional cleanup (Section 11).

15.    METHOD PERFORMANCE

15.1    The method detection  limit (MDL) is  defined as the minimum concentration of a
       substance that can be measured and reported with 99% confidence that the value is above
       zero.13 The MDL concentrations listed in Table 1 were obtained using reagent watet.

15.2    In a single laboratory, Susquehanna University, using spiked tap water samples,  the
       average recoveries presented in Table 3 were obtained.  The standard deviation of the
       percent recovery is also included in Table 3.14

-------
Method 614
                                    References

1.      "Methods  for  Benzidine, Chlorinated  Organic Compounds, Pentachlorophenol  and
       Pesticides   in  Water  and  Wastewater,"   U.S.  Environmental Protection  Agency,
       Environmental Monitoring and Support Laboratory - Cincinnati, Ohio, September 1978.

2.      ASTM Annual Book of Standards, Part 31, D3694, "Standard Practice for Preparation of
       Sample Containers and for  Preservation," American Society for Testing and Materials,
       Philadelphia, PA, p. 679, 1980.

3.      "Carcinogens - Working with Carcinogens," Department of Health,  Education,  and
       Welfare, Public  Health  Service, Center for  Disease  Control,  National  Institute  for
       Occupational Safety and Health, Publication No. 77-206, August 1977.

4.      "OSHA Safety and Health Standards, General Industry," (29 CFR 1910), Occupational
       Safety and Health Administration, OSHA 2206, (Revised, January 1976).

5.      "Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
       Committee on  Chemical Safety, 3rd Edition, 1979.

6.      ASTM Annual Book of  Standards, Part 31, D3086, Appendix X3, "Standardization of
       Florisil Column by Weight Adjustment Based on Adsorption of Laurie Acid," American
       Society for Testing and Materials, Philadelphia, PA, p 765,  1980.

7.      "Handbook for Analytical  Quality Control  in Water and Wastewater Laboratories,"
       EPA-600/4-79-019, U. S.  Environmental Protection Agency, Environmental Monitoring
       and Support Laboratory - Cincinnati, Ohio,  March 1979.

8.      ASTM Annual  Book of Standards, Part 31, D3370, "Standard Practice for Sampling Water,"
       American Society for Testing and Materials, Philadelphia, PA, p. 76,  1980.

9.      Law, L. M. and D. F. Goerlitz. "Microcolumn Chromatographic Cleanup for the  Analysis
       of Pesticides in Water," Journal of the Association of Official Analytical  Chemists,  53, 1276,
       (1970).

10.     Burke, J. A.   "Gas Chromatography for Pesticide Residue  Analysis; Some  Practical
       Aspects," Journal of the Association of Official Analytical Chemists, 48, 1037 (1965).

11.     McNair, H.M.  and Bonelli, E. J. Basic Chromatography, Consolidated Printing, Berkeley,
       California, p. 52, 1969.

12.     Eichelberger, J.W., Harris, L.E., and Budde, W.L. "Reference Compound to Calibrate Ion
       Abundance  Measurement  in  Gas  Chromatography-Mass  Spectrometry," Analytical
       Chemistry,  47, 995 (1975).

13.     Glaser, J.A. et.al, "Trace Analysis for Wastewaters," Environmental Science & Technology,
       15, 1426 (1981).

-------
                                                                                Method 614
14.     McGrath, T. F., "Recovery Studies of Pesticides From Surface and Drinking Waters," Final
       Report for U.S. EPA Grant R804294, Environmental Monitoring and Support Laboratory,
       Cincinnati, Ohio.

-------
Method 614
Table 1.      Chromatographic Conditions and Method Detection Limits
                  Parameter
                  Diazinon
                  Disulfoton
                  Demeton
                  Parathion-
                  methyl
                  Malathion
                  Parathion-
                  ethyl
                  Ethion
                  Azinphos-
                  methyl
                                      Retention Time
                                           (min)
Column 1

     1.8
     1.9
     2.3
     2.5

     2.9
     3.1

     6.8
    14.5
Column 2

     1.8
     2.1
     2.1
     3.7

     3.9
     4.5

     9.1
    29.9
                          Method
                         Detection
                           Limit
0.012
 ND
 ND
0.012

 ND
0.012

 ND
 ND
ND = Not determined

Column 1 conditions: Gas-Chrom Q (100/120 mesh) coated with 3% OV-1 packed in a glass
column 1.8 m long by 4 mm ID with nitrogen carrier gas at a flow rate of 60 mL/min.  Column
temperature, isothermal at 200°C.  A flame photometric detector was used with this column to
determine the MDL.

Column 2 conditions:  Gas Chrom  Q (100/120 mesh) coated with 1.5% OV-17/1.95% QF-1
packed in a glass column 1.8 m long by 4 mm ID with nitrogen carrier  gas at a flow rate of
70 mL/min. Column temperature, isothermal at 212°C.

-------
                                                                                Method 614
Table 2.
Single-Operator Accuracy and Precision
       Parameter
       Diazinon
       Parathion
       methyl
       Parathion ethyl    102
Average
Percent
Recovery
94
95
Standard
Deviation
(%)
5.2
3.2
Spike
Range
(Vg/L)
0.04-40
0.06-60
Number of
Analyses
27
27
Matrix
Types
4
4
                          4.1
         0.07-70
              27
Table 3.
Florisil Fractionation Patterns and Acetonitrile Partition Applicability
                                Percent Recovery by Fraction
 Parameter
 Demeton
 Disulfoton
 Diazinon
 Malathion
 Parathion ethyl
 Parathion methyl
 Azinphos methyl
 Ethion

ND = Not  determined
No. 1
No. 2
No. 3
No. 4
100
100
              ND
100
  5
100
100

ND
                                       95
 20
ND
Florisil eluate composition by fraction:
       Fraction 1 = 200 mL of 6% ethyl ether in hexane
       Fraction 2 = 200 mL of 15% ethyl ether in hexane
       Fraction 3 = 200 mL of 50% ethyl ether in hexane
       Fraction 4 = 200 mL of ethyl ether
 80
ND
 Acetonitrile
  Partition
Applicability
     ND
     ND
     Yes
     Yes
     Yes
     Yes
     ND
     Yes

-------