SEPA
United States
Environmental Protection
Agency
 www.epa.gov
  Method 614.1: The Determination
  of Organophosphorus Pesticides
  in Municipal and Industrial
  Waste water

-------
             Method 614.1
         The Determination of
Organophosphorus Pesticides in
      Municipal and Industrial
                  Wastewater

-------
                                Method 614.1
 The Determination of Organophosphorus Pesticides in Municipal and
                            Industrial Wastewater


1.     SCOPE AND APPLICATION

1.1    This method covers the  determination  of  certain  organophosphorus pesticides in
      municipal and industrial wastewater. The following parameters may be determined by
      this method.

               Parameter             STORET No.          CAS No.
               Dioxathion                                   78-34-2
               EPN                                      2104-64-5
               Ethion                   39398              563-12-2
               Terbufos                                  13071-79-9

1.2    The estimated detection limit (EDL) for each parameter is listed in Table 1. The EDL was
      calculated from the minimum detectable response of the nitrogen/phosphorus detector
      equal to 5  times the gas chromatographic (GC) background noise assuming a 1.0-mL
      final extract volume of a 1-L reagent water sample and an injection of 5 uL.  The EDL
      for a specific wastewater may be different depending on the nature of interferences in
      the sample matrix.

1.3    This is a  gas chromatographic (GC) method applicable to  the determination of the
      compounds listed above in municipal and industrial  discharges. When this method is
      used to analyze  unfamiliar samples for  any or  all  of the compounds listed above,
      compound identifications should be supported by at least one additional qualitative
      technique.   Section  13 provides gas  chromatograph/mass spectrometer  (GC/MS)
      conditions appropriate for the qualitative confirmation of compound identifications.

1.4    This method is restricted to use by or under the supervision of analysts experienced in
      the operation of gas chromatographs and in the interpretation of chromatograms.

2.     SUMMARY OF METHOD

2.1    Organophosphorus pesticides are removed from the  sample matrix by extraction with
      15%  methylene chloride in hexane.  The extract is dried, exchanged into hexane, and
      analyzed by gas chromatography.  Column chromatography is used as necessary to
      eliminate interferences which may be encountered.  Measurement of the pesticides is
      accomplished with a nitrogen/phosphorus-specific detector.

2.2    Confirmatory analysis by  GC/MS is recommended when a new or undefined sample
      type is being analyzed if the concentration is adequate for such determination.

-------
Method 614-1
3.     INTERFERENCES

3.1     Solvent, reagents, glassware, and other sample-processing hardware may yield discrete
       artifacts and/or elevated baselines causing misinterpretation of gas chromatograms. All
       of these materials must be  demonstrated  to  be free  from interferences under the
       conditions of the analysis by running laboratory reagent blanks as described in Section
       9.1.

       3.1.1   The use of high-purity reagents  and solvents helps to minimize interference
              problems.   Purification of solvents by distillation in all-glass systems may be
              required.

       3.1.2   Glassware must be scrupulously cleaned.1 Clean all glassware as soon as possible
              after use by rinsing with the last solvent used in it. This should be followed by
              detergent washing with hot water and rinses with tap water and reagent water.
              It should then be drained dry and heated in a muffle furnace at 400°C for  15 to
              30 minutes.  Solvent rinses with acetone and pesticide-quality hexane may be
              substituted for the  heating.  Volumetric ware should not be heated in a muffle
              furnace. After drying and cooling, glassware should be sealed and stored in a
              clean environment to prevent any accumulation of dust or other contaminants.
              Store the glassware inverted or  capped with aluminum foil.

3.2     Interferences coextracted from the samples will vary considerably from source to source,
       depending  on the diversity of the industrial complex or municipality being sampled.
       While general cleanup procedures are provided as part of this method, unique samples
       may require  additional cleanup approaches to achieve  the detection limits listed in
       Table 1.

4.     SAFETY

4.1     The toxicity or carcinogenicity of each reagent used in this method has not been precisely
       defined;  however, each chemical compound should be  treated as a potential health
       hazard. From this viewpoint,  exposure  to these chemicals must be reduced to the lowest
       possible level by whatever means available. The laboratory is responsible for maintaining
       a current awareness file of  OSHA regulations  regarding the safe  handling of the
       chemicals specified in this method.  A reference  file of material data handling sheets
       should  also be made  available to  all personnel involved in  the  chemical  analysis.
       Additional references to laboratory safety are  available and have been identified2 4 for the
       information of the analyst.

5.     APPARATUS AND EQUIPMENT

5.1     Sample containers: Narrow-mouth glass  bottles, 1-L or 1-quart volume, equipped with
       polytetrafluoroethylene  (PTFE)-lined screw-caps.  Wide-mouth glass bottles, 1-quart
       volume, equipped with PTFE-lined screw-caps may also be used.   Prior to use, wash
       bottles and cap liners with detergent and rinse with tap and distilled water. Allow the
       bottles and  cap liners to air dry, then muffle at 400°C for 1 hour.  After cooling, rinse the
       cap liners with hexane, seal the bottles, and store  in a dust-free environment.

-------
                                                                              Method 614-1
       5.1.1  Automatic sampler (optional):  Must incorporate glass sample containers for the
             collection of a minimum of 250 mL. Sample containers must be kept refrigerated
             at 4°C  and protected from light  during compositing.   If the sampler uses a
             peristaltic pump, a minimum length of compressible silicone rubber tubing may
             be used.  Before use, however, the compressible tubing should  be thoroughly
             rinsed with methanol,  followed by repeated rinsings  with distilled  water to
             minimize the potential for contamination  of the sample.  An integrating  flow
             meter is required to collect flow-proportional composites.

5.2     Kuderna-Danish (K-D)  glassware.

       5.2.1  Synder column:  Three-ball  macro (Kontes K-503000-0121  or equivalent) and
             two-ball micro (Kontes  K-569001-0219  or equivalent).

       5.2.2  Concentrator tube:  10-mL, graduated (Kontes K-570050-1025 or equivalent)  with
             ground-glass stopper.

       5.2.3  Evaporative flask:   500-mL (Kontes K-570001-0500 or  equivalent).  Attach to
             concentrator tube with springs.

5.3     Gas chromatography system.

       5.3.1  The gas chromatograph  must be  equipped with  a glass-lined  injection  port
             compatible with the detector  to be used.  A data  system is recommended for
             measuring peak areas.

             5.3.1.1 Column  1:  180 cm long by 2 mm ID, glass, packed with  3% OV-225 on
                    Supelcoport (100/120 mesh) or equivalent.

             5.3.1.2 Column  2:   120  cm  long by 2 mm  ID,  PyrexR glass,  packed  with
                    1.5% OV-17/1.95 % QF-1 on Gas Chrom Q,  80/100 mesh or equivalent.

             5.3.1.3 Column  1 was used to develop the  accuracy and precision statements in
                    Section 12.  Guidelines for the use of alternative column packings are
                    provided in Section 10.3.1.

             5.3.1.4 Detector: nitrogen/phosphorus. This detector has proven effective in the
                    analysis  of wastewaters for the parameters  listed in the scope  and was
                    used to  develop  the  method  performance statements in Section 12.
                    Guidelines  for  the  use  of  alternative  detectors  are   provided  in
                    Section 10.3.1.

5.4     Chromatographic column:  300  mm long by 10  mm ID  Chromaflex, equipped  with
       coarse-fitted bottom plate and PTFE stopcock. (Kontes K-420540-0213 or equivalent).

5.5     Drying column: Approximately 400 mm long by 20 mm ID borosilicate glass, equipped
       with coarse-fitted bottom plate.

-------
Method 614-1
5.6     Miscellaneous.

       5.6.1  Balance: Analytical, capable of accurately weighing to the nearest 0.0001 g.

       5.6.2  Separatory funnel:  2-L, equipped with PTFE stopcock.

       5.6.3  Water bath:  Heated with concentric ring cover, capable of temperature control
             (±2°C). The bath should be used in a hood.

       5.6.4  Standard solution storage containers:  15-mL bottles with PTFE-lined screw-caps.

       5.6.5  Boiling chips:  Approximately 10/40 mesh.  Heat to 400°C for 30 minutes  or
             perform a Soxhlet extraction with methylene chloride.

6.     REAGENTS AND CONSUMABLE MATERIALS

6.1     Reagents.

       6.1.1  Acetone, hexane, and methylene chloride: Demonstrated to be free of analytes.

       6.1.2  Silica gel:  Woelm 70-230 mesh.  Activate approximately 100 g of silica gel at
             200°C for 6 hours in a tared 500-mL Erlenmeyer flask with ground-glass stopper.
             Allow to cool to room temperature, reweigh, and determine the weight  of
             activated silica  gel.  Deactivate  by adding  3% by  weight of distilled water.
             Restopper the flask, and shake on a wrist-action shaker for at least one hour.
             Allow to equilibrate for 3 or more hours at room temperature.

       6.1.3  Reagent water: Reagent water is defined as a water in which an interferent is not
             observed at the  method detection limit of each  parameter of interest.

       6.1.4  Sodium hydroxide (NaOH) solution (ION): Dissolve 40 g NaOH in reagent water
             and dilute to 100 mL.

       6.1.5  Sodium sulfate:  Granular, anhydrous. Condition by heating  at 400°C for 4 hours
             in a shallow tray.
       6.1.6  Sulfuric acid  (HjSOJ solution (1 + 1):  Add measured volume of concentrated
             H2SO4 to equal volume of reagent water.

6.2     Standard stock solutions (1.00  ug/uL): These solutions may be purchased as certified
       solutions or prepared from pure standard materials using the following procedures.

       6.2.1  Prepare standard stock solutions by  accurately weighing about 0.0100 g of pure
             material.  Dissolve the material in  hexane and dilute to volume in  a 10-mL
             volumetric flask. Larger volumes can be used at the convenience of the analyst.
             If compound purity is certified at 96% or greater, the weight can be used without
             correction to calculate the concentration of the standard stock.

-------
                                                                              Method 614-1
       6.2.2   Store standard stock solutions at 4°C in 15-mL bottles equipped with PTFE-lined
              screw-caps. Standard stock solutions should be checked frequently for signs of
              degradation or  evaporation,  especially just  prior to preparing calibration
              standards from them.

       6.2.3   Standard stock solutions must be replaced after 6 months, or sooner if comparison
              with check standards indicates a problem.

7.     SAMPLE COLLECTION, PRESERVATION, AND STORAGE

7.1     Collect all samples in duplicate.  Grab samples must be collected in glass containers.
       Conventional sampling practices5 should be followed, except that the bottle must not be
       prewashed with sample before collection.

7.2     The samples must be iced  or  refrigerated at 4°C from  the  time of collection  until
       extraction.

7.3     Chemical preservatives should not be used in the field unless  more than 24 hours will
       elapse  before delivery to the laboratory.  If the samples will  not be  extracted within
       48 hours of collection, the sample should be adjusted to a pH range of 6.0 to  8.0 with
       sodium hydroxide  or  sulfuric acid.

7.4     All samples must be extracted within 7 days and completely analyzed within 40 days of
       extraction.6

8.     CALIBRATION AND STANDARDIZATION

8.1     Calibration.

       8.1.1   A set of at least three calibration solutions containing the method analytes is
              needed.  One calibration solution should contain each analyte at a concentration
              approaching but greater than  the estimated detection  limit (Table 1)  for that
              compound; the other two solutions should contain analytes at concentrations that
              bracket the range expected in samples. For example, if the detection limit for a
              particular analyte is 0.2 l-ig/L, and a sample expected to contain approximately
              5 l-ig/L is analyzed, solutions of standards should be prepared at concentrations
              of 0.3 l-ig/L, 5  l-ig/L, and 10 l-ig/L for the particular analyte.

       8.1.2   To prepare a calibration solution, add an appropriate volume of a standard stock
              solution to a volumetric flask and dilute to volume with hexane.

       8.1.3   Starting  with  the standard  of lowest concentration, analyze  each calibration
              standard according to Section 10.3.2 and tabulate peak height or area responses
              versus the  mass of analyte injected.   The results can be used to prepare  a
              calibration curve for each compound.   Alternatively, if the ratio of response to
              concentration  (calibration factor) is  a  constant  over  the working range (<10%
              relative standard deviation), linearity through the origin can be assumed and the
              average  ratio or calibration factor can be used in place of a calibration curve.

-------
Method 614-1
       8.1.4   The working calibration curve or calibration factor must be verified on each
              working day by the measurement of one or more calibration standards.  If the
              response for any analyte varies from the predicted response by more than ±10%,
              the test must be repeated using a fresh calibration standard. If the results still do
              not agree, generate a new calibration curve.

9.     QUALITY CONTROL

9.1     Monitoring for interferences:   Analyze a laboratory reagent blank  each time a set  of
       samples is extracted. A laboratory reagent blank is a 1-L aliquot of reagent water. If the
       reagent blank contains a reportable level of any analyte, immediately check the entire
       analytical system to locate and correct for possible  interferences and repeat the test.

9.2     Assessing accuracy.

       9.2.1   After  every 10  samples, and preferably in  the  middle of each day, analyze a
              laboratory control standard. Calibration standards may not be used for accuracy
              assessments and the laboratory control standard may not be used for calibration
              of the analytical system.

              9.2.1.1 Laboratory control standard concentrate:  From stock standards prepared
                    as  described in Section  6.2, prepare  a  laboratory control  standard
                    concentrate  that contains  each analyte of interest at  a concentration  of
                    2 ug/mL in acetone or other  suitable solvent.7

              9.2.1.2 Laboratory  control standard:  Using a pipette,  add  1.00  mL of the
                    laboratory control standard concentrate to a 1-L aliquot of reagent water.

              9.2.1.3 Analyze the laboratory control standard as described in Section 10.  For
                    each analyte in the laboratory control  standard, calculate the percent
                    recovery (Pj) with the equation:


                                        Equation 1

                                             1005.
                                        r,
              where
              St  = Analytical results from the laboratory control standard, in ug/L
              Tt  = Known concentration of the spike, in ug/L
       9.2.2   At least annually,  the laboratory should  participate in formal performance
              evaluation studies, where solutions of unknown concentrations are analyzed and
              the performance of all participants is compared.

9.3     Assessing precision.

-------
                                                                               Method 614-1
       9.3.1  Precision assessments for this  method are based  upon  the  analysis of field
             duplicates (Section 7.1).  Analyze both sample bottles for  at least  10% of all
             samples.  To the extent practical, the  samples for  duplication should contain
             reportable levels of most of the  analytes.

       9.3.2  For each analyte in each duplicate pair, calculate the relative range7 (RRj) with the
             equation:
              where
              Rt  = Absolute difference between the duplicate measurements %  and £ ,  in ug/L

                                            ( X. + X7\
              Xt  = Average concentration found   —=	=  , in ug/L
       9.3.3  Individual relative range measurements are pooled to determine average relative
             range or to develop an expression of relative range as a function of concentration.

10.    PROCEDURE

10.1    Sample extraction.

       10.1.1 Mark the water meniscus on the side of the sample bottle for later determination
             of sample  volume.  Pour the entire sample into a 2-L separatory funnel.  Check
             the pH of  the sample with wide-range pH paper and adjust to within the range
             of 5 to 9 with sodium hydroxide or sulfuric acid.

       10.1.2 Add 60 mL of 15% methylene chloride/hexane to the sample bottle and shake for
             30 seconds to rinse the walls. Transfer the solvent to the separatory funnel and
             extract the sample by shaking the funnel for 2 minutes with periodic venting to
             release vapor pressure.  Allow the organic layer to separate from the water phase
             for a minimum of 10 minutes.  If the emulsion interface between layers is more
             than one-third the  volume of the solvent  layer,  the  analyst  must  employ
             mechanical techniques to complete the phase separation.  The optimum technique
             depends on  the  sample, but may  include stirring, filtration of the emulsion
             through glass wool, or centrifugation. Collect the extract in a 250-mL Erlenmeyer
             flask.

       10.1.3 Add an additional 60-mL  volume  of 15% methylene chloride/hexane to  the
             sample bottle and complete the extraction procedure a second time, combining the
             extracts in the  Erlenmeyer flask.

-------
Method 614-1
       10.1.4  Perform a third extraction in  the same manner.  Pour the combined extract
              through a drying column containing about 10 cm of anhydrous sodium sulfate,
              and collect it in a 500-mL K-D flask equipped with a 10-mL concentrator tube.
              Rinse the Erlenmeyer flask and column with 20 to 30 mL of methylene chloride
              to complete the quantitative transfer.

       10.1.5  Add one or two clean boiling  chips to the flask and attach a three-ball Snyder
              column. Prewet the Snyder column by adding about 1 mL of methylene chloride
              to the top.  Place the K-D apparatus on a hot water bath (80 to 85°C) so that the
              concentrator tube is partially immersed in the hot water and the entire lower
              rounded surface of the flask is bathed in steam.  Adjust the vertical position of
              the  apparatus  and  the water  temperature as  required to  complete  the
              concentration in 15 to 20 minutes. At the proper rate of distillation,  the balls of
              the  column will actively chatter but the chambers  will not flood.  When the
              apparent volume of liquid reaches 1 mL, remove the  K-D apparatus and allow it
              to drain and cool for at least 10 minutes.  Remove the Snyder column and rinse
              the flask and its lower joint into the concentrator tube with 1 to 2 mL of hexane.
              A 5-mL syringe is recommended for  this operation.  If the  extract requires
              cleanup, proceed to Section 10.2  (cleanup and separation). If cleanup has been
              performed or if the extract does not require cleanup, proceed with Section 10.1.6.

       10.1.6  Add a clean boiling chip to the concentrator tube.  Attach a two-ball micro-Snyder
              column. Prewet the micro-Snyder column by adding about 0.5 mL of hexane to
              the top. Place  this micro K-D  apparatus on a steaming-water bath (80 to 85°C)
              so that the concentrator tube is partially immersed in the hot water.  Adjust the
              vertical position of the apparatus  and water temperature as required to complete
              the concentration in 5 to 10 minutes.  At the proper rate of distillation, the balls
              will actively chatter but the chambers will not flood.  When the apparent volume
              of liquid reaches 0.5 mL, remove the K-D apparatus and allow it to drain and cool
              for at least 10  minutes.  Remove the micro-Snyder column and rinse its lower
              joint into the concentrator tube with a small volume  of hexane.  Adjust the final
              volume to 1.0 mL or to a volume suitable for cleanup or gas chromatography, and
              stopper the concentrator tube; store refrigerated if further processing will not be
              performed immediately. If the  extracts will be stored longer than two days, they
              should be transferred  to  PTFE-sealed screw-cap bottles.   Proceed with  gas
              chromatographic analysis.

       10.1.7  Determine the  original sample volume by refilling the sample bottle to the mark
              and transferring the liquid to a 1000-mL graduated cylinder.  Record the sample
              volume to the nearest 5 mL.

10.2    Cleanup and separation.

       10.2.1  Cleanup procedures may not be necessary for a relatively clean sample matrix.
              The cleanup procedures recommended in this method have been used for the
              analysis of various clean waters and municipal effluents. The silica gel procedure
              allows for a select fractionation of the  compounds and will eliminate non-polar
              materials.  The single-operator precision and accuracy data in Table 2  were
              gathered using  the recommended cleanup procedures.  If particular circumstances
              demand the use of an alternative  cleanup procedure,  the analyst must determine

-------
                                                                              Method 614-1
             the elution  profile and demonstrate that the recovery of each compound of
             interest is no less than that recorded in Table 2.

       10.2.2 Prepare silica gel  columns using a glass column 200 mm long by 10 mm ID.
             Rinse column with hexane.  Add approximately 50 mL of hexane to the empty
             column.  Add 3.5 grams of 3% deactivated silica gel. Pack by rotating slowly to
             release air bubbles.  Top with 1.5 cm of Na2SO4.  Drain hexane to the top of
             Na2SO4 layer.

       10.2.3 Just prior to exposure of the sodium sulfate  layer to the air, transfer the sample
             extract onto the column using an additional 2 mL of hexane  to complete the
             transfer.

       10.2.4 Just prior to exposure of the sodium sulfate layer to the air, add 30 mL of 6%
             methylene chloride/hexane and continue the elution of the column, collecting the
             eluate in a 500-mL K-D flask equipped with a 10-mL concentration tube.  Elution
             of the column should be  at a rate of about 2 mL per minute. Add 50 mL of
             hexane to the flask  and concentrate the  collected fraction  by  the standard
             technique prescribed in Sections 10.1.5 and 10.1.6.

       10.2.5 Continue the elution of the column according to the scheme outlined in Table 3.
             The elution of the compounds may vary with different sample matrices.

       10.2.6 Analyze the fractions by gas chromatography.

10.3    Gas chromatographic analysis.

       10.3.1 Recommended columns  and detector for the gas chromatography system are
             described in Section  5.3.1.  Table 1 summarizes the recommended  operating
             conditions for the gas chromatograph.  Included  in  this table are  estimated
             retention  times  and  detection limits that  can  be achieved  by  this  method.
             Examples of the separations achieved are shown in Figures 1 and  2.  Other
             packed columns, chromatographic conditions, or detectors may be used if data
             quality comparable to Table 2 are achieved. Capillary (open-tubular) columns
             may also be used if the relative standard deviations of responses for replicate
             injections are  demonstrated to be less than  6% and data quality comparable to
             Table 2 are  achieved.

       10.3.2 Inject 2 to 5 uL of the sample extract using the solvent-flush technique.8 Record
             the volume injected to the nearest 0.05 uL, the total  extract volume, and the
             resulting peak size in area or peak height units.

       10.3.3 The width of the retention-time window used to make identifications should be
             based upon measurements of actual  retention-time variations of standards over
             the course of the day. Three times the standard deviation of a retention time for
             a compound can be used to calculate a suggested window size; however, the
             experience  of the analyst should  weigh  heavily  in  the interpretation of
             chromatograms.

-------
Method 614-1
       10.3.4 If the response for the peak exceeds the working range of the system, dilute the
             extract and reanalyze.

       10.3.5 If the measurement of the peak  response is  prevented  by the presence of
             interferences, further cleanup is required.

11.    CALCULATIONS

11.1    Determine the concentration (C) of individual compounds in the sample in micrograms
       per liter with the equation:


                                      Equation 3

                                     c  =  (A)  (V)
             where
             A  = Amount of material injected, in ng
             Vt  = Volume of extract injected, in uL
             Vt  = Volume of total extract, in uL
             V,  = Volume of water extracted, in ml
11.2    Report the results for the unknown samples in microgram per liter. Round off the results
       to the nearest 0.1 l-ig/L or two significant figures.

12.    METHOD PERFORMANCE

12.1    Estimated detection limits (EDL)  and associated chromatographic conditions are listed
       in Table I.9 The detection limits were calculated from the minimum detectable response
       of the N/P detector equal to 5 times the GC background noise, assuming a  1.0 mL final
       extract volume of a 1-L sample and a GC injection of 5 uL.
12.2    Single laboratory accuracy and precision studies were conducted by ESE,6 using spiked
       relevant industrial wastewater samples.  The results of these studies are presented in
       Table 2.

13.    GC/MS CONFIRMATION

13.1    It is recommended that GC/MS  techniques  be judiciously  employed to support
       qualitative identifications made  with this method.  The mass spectrometer should be
       capable of scanning the mass range from 35 amu to a mass 50 amu above the molecular
       weight of the compounds of interest.  The instrument must be capable of scanning the
       mass range at a rate to produce at least 5 scans per peak, but not to exceed 7 scans per
       peak utilizing a 70 V (nominal) electron energy in the electron impact ionization mode.
       A GC-to-MS interface constructed of all-glass or glass-lined materials is recommended.

-------
                                                                              Method 614-1
       A computer system  should be interfaced to the mass spectrometer that allows the
       continuous acquisition and  storage on  machine-readable media of all mass spectra
       obtained throughout  the duration of the chromatographic program.

13.2    Gas chromatographic  columns and conditions should be selected for optimum separation
       and performance. The conditions selected must be  compatible with standard GC/MS
       operating practices.  Chromatographic tailing factors of less than  5.0 must be achieved.
       The calculation of tailing factors is illustrated in Method 625.10

13.3    At the beginning  of each day that confirmatory analyses are to be performed, the GC/MS
       system must be checked to see that all DFTPP performance criteria are achieved.11

13.4    To confirm an identification of a compound, the background-corrected mass spectrum
       of the compound must be obtained from the  sample extract and compared with a mass
       spectrum from a  stock or calibration standard analyzed under the same chromatographic
       conditions.  It is recommended that at least 25 ng of material be injected into the GC/MS.
       The criteria below must be met for qualitative confirmation.

       13.4.1 The molecular ion and other ions that are present above 10% relative abundance
             in the mass spectrum of the standard must be  present in the mass spectrum of the
             sample with agreement to ±10%. For example, if the relative abundance of an ion
             is 30% in  the mass spectrum of the standard,  the allowable limits for the relative
             abundance of  that ion in the mass spectrum for the sample would be 20 to 40%.

       13.4.2 The retention-time of the compound in the sample must be within 6 seconds  of
             the same  compound in the standard solution.

       13.4.3 Compounds that have similar mass spectra can be explicitly identified by GC/MS
             only on the basis of retention-time data.

13.5    Where available, chemical  ionization  mass  spectra may be employed to aid in the
       qualitative identification process.

13.6    Should these MS procedures fail to provide satisfactory results, additional steps may  be
       taken before reanalysis. These may include the use of alternative packed or capillary GC
       columns or additional cleanup.

-------
Method 614-1
                                    References

1.      ASTM Annual Book of Standards, Part 31, D3694, "Standard Practice for Preparation of
       Sample Containers and for Preservation," American Society for Testing and Materials,
       Philadelphia, PA, p. 679, 1980.

2.      "Carcinogens—Working  with  Carcinogens," Department of Health, Education, and
       Welfare,  Public Health  Service, Center for Disease  Control, National Institute for
       Occupational Safety and Health, Publication No. 77-206, August 1977.

3.      "OSHA Safety and Health Standards, General  Industry" (29 CFR 1910), Occupational
       Safety and Health Administration, OSHA 2206  (Revised, January  1976).

4.      "Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
       Committee on  Chemical  Safety, 3rd Edition, 1979.

5.      ASTM  Annual Book of Standards,  Part 31, D3370, "Standard Practice for Sampling
       Water," American Society for Testing and Materials, Philadelphia, PA, p. 76, 1980.

6.      Test  procedures for Pesticides in Wastewaters,  EPA Contract Report #68-03-2897.
       Unpublished report available from U.S. Environmental Protection Agency, Environmental
       Monitoring and Support Laboratory, Cincinnati, Ohio.

7.      "Handbook for Analytical Quality Control in Water  and Wastewater Laboratories,"
       EPA-600/4-79-019, U.S. Environmental Protection Agency, Environmental Monitoring and
       Support Laboratory Cincinnati, Ohio  45268, March 1979.

8.      Burke,  J.A.  "Gas Chromatography  for Pesticide Residue Analysis;  Some Practical
       Aspects," Journal of the Association of Official Analytical Chemists, 48, 1037  (1965T)

9.      "Evaluation of  Ten Pesticide Methods," U.S. Environmental Protection Agency, Contract
       No. 68-03-1760, Task No. 11, U.S. Environmental Monitoring and Support Laboratory,
       Cincinnati, Ohio  45268.

10.     "Methods for  Organic Chemical Analysis of Municipal and Industrial Wastewater,"
       EPA-600/4-82-057, U.S. Environmental Protection Agency, Environmental Monitoring and
       Support Laboratory, Cincinnati, Ohio 45268.

11.     Eichelberger, J.W., Harris, L.E., and Budde, W.L.  Analytical Chemistry, 46, 1912 (1975).

-------
                                                                             Method 614-1
Table 1.      Gas Chromatography of Organophosphorus Pesticides

                                Retention Time (min)              Detection Limit
Parameter                  Column 1     |     Column 2                (l*g/L)
Terbufos                     1.41                    1.9                      .004
Dioxathion                   2.3                     2.3                      .01
Ethion                       8.3                     6.4                     0.1
EPN                        13.3                     8.3                     0.2

Column 1: 180 cm long by 2 mm ID, glass, packed with 3% OV-225 on 100/120 Supelcoport;
nitrogen carrier gas at a flow rate of 50 mL/min.  Column temperature is 200°C for 2 minutes,
then programmed at 5°/min. to 240°C and held for 5 minutes.

Column 2: 120 cm long by 2 mm ID, PyrexR glass,  packed with 1.5% OC-17/1.95% QF-1 on
80/100 mesh Gas Chrom Q or equivalent; nitrogen  carrier gas at a flow rate  of 30 mL/min.
Column temperature  is 180°C for 2 minutes, then programmed at 8°/min. to 250°C and held for
4 minutes.
Table 2.      Single-Laboratory Accuracy and Precision

                   Matrix  Spike Range    Number of   Average Percent   Standard
  Parameter        Type*      fag/L)       Replicates       Recovery    Deviation (%)
  Dioxathion            1        1,978.0           7              94.3          19.9
                        1           19.8           7              99.0          27.5

  EPN                  1        1,293.0           7              96.1          6.1

  Ethion                1        1,788.0           7              89.2          4.5

  Terbufos              1           15.1           7             101.0          12.4
                        1        1,508.0           7              95.0          3.4

*1 = Combined industrial wastewaters

-------
Method 614-1
Table 3. Silica Gel Cleanup of Organophosphorus Pesticides
Percent Recoveries
Silica Gel Fraction* Terbufos
1 0
2 0
3 93.0
4 0
Dioxathion Ethion
0 0.8
0 1.9
35.1 94.9
52.7 3.0
EPN
0
0
46.4
56.0
  Total Percent Recoveries         93.0           87.8           101               102
*Fraction 1 = 30 mL 6% MeCl2 in hexane
 Fraction 2 = 30 mL 15% MeCl2 in hexane
 Fraction 3 = 30 mL 50% MeCl2 in hexane
 Fraction 4 = 30 mL 100% MeCl2

-------
                                                                              Method 614-1
                Terbufus
                                           Ethion
                  Dioxathion
                                                              EPN
                    i    i    i—i—i—i—i—i—i—i—r~~\—r~
                   4.0     6.0     8.0    10.0    12.0     14.0    16.0
                            Retention Time (minutes)
                                                                         AS2-002-37A
Figure 1.   Gas Chromatogram of Organophosphorous Pesticides (Column 1)

-------
Method 614-1
                                                       Ethion
                                                                 EPN
                 —i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i—i
                  0   1.0  2.0  3.0  4.0  5.0   6.0   7.0   8.0   9.0   10.0  11.0  12.0
                                  Retention Time (minutes)
                                                                                A52-002-38A
        Figure 2.  Gas Chromatogram of Organophosphorous Pesticides (Column 2)

-------