SEPA
United States
Environmental Protection
Agency
 www.epa.gov
   Method 615: The Determination of
   Chlorinated Herbicides in
   Municipal and Industrial
   Waste water

-------
         Method 615
    The Determination of
Chlorinated Herbicides in
Municipal and Industrial
            Wastewater

-------
                                 Method 615
     The Determination of Chlorinated Herbicides in Municipal and
                             Industrial Wastewater


1.     SCOPE AND APPLICATION

1.1    This method covers the determination of certain chlorinated herbicides. The following
      parent acids can be determined by this method:

                     Parameter     STORET No.    CAS No.
                     2,4-D              39736       94-75-7
                     Dalapon              -         75-99-0
                     2,4-DB               -         94-82-6
                     Dicamba             -        1918-00-9
                     Dichlorprop           -         120-36-5
                     Dinoseb              -         88-85-7
                     MCPA               -         94-74-6
                     MCPP               -        7085-19-0
                     2,4,5-T            39740       93-76-5
                     2,4,5-TP           39760       93-72-1

1.2    This method  is also  applicable to the determination of salts and  esters of these
      compounds.  These include,  but are not limited to: the isobutyl and isooctyl esters of
      2,4-D; the isobutyl and isooctyl esters of 2,4-DB; the isooctyl ester of MCPA; and the
      isooctyl ester of 2,4,5-TP.  The actual form of each acid is not distinguished by  this
      method. Results are calculated and reported for each listed parameter as total free acid.

1.3    This is  a gas chromatographic (GC) method applicable to the determination of the
      compounds listed  above in  industrial and municipal discharges as provided under
      40 CFR 136.1.  Any modification of this method beyond those expressly permitted shall
      be considered a major  modification subject to application and approval of alternative test
      procedures under 40 CFR 136.4 and  136.5.

1.4    The method detection  limit (MDL, defined in Section 15) for each parameter is listed in
      Table 1. The MDL for a specific wastewater may differ from  those listed, depending
      upon the nature of interferences in the sample matrix.

1.5    This method is restricted to use by or under the supervision of analysts experienced in
      the use of gas chromatography and  in the interpretation of gas chromatograms. Each
      analyst must demonstrate the  ability to generate acceptable results with this method
      using the procedure described  in Section 8.2.

1.6    When this method  is used to analyze unfamiliar samples for any or all of the compounds
      above,  compound  identifications  should  be supported by  at least  one  additional
      qualitative technique.  This method describes analytical conditions for alternative  gas
      chromatographic columns that can be used to confirm measurements made with the

-------
Method 615
       primary column. Section 15 provides gas chromatograph/ mass spectrometer (GC/MS)
       criteria appropriate for the  qualitative confirmation of compound identifications.

2.     SUMMARY OF METHOD

2.1     A measured volume of sample, approximately 1 L, is acidified. The acid herbicides and
       their esters and salts are extracted with ethyl ether using a separatory funnel.  The
       derivatives are hydrolyzed with potassium hydroxide and extraneous organic material
       is removed by a solvent wash.  After acidification, the acids are extracted and converted
       to their methyl esters using diazomethane as the derivatizing agent.  Excess reagent is
       removed, and the esters are determined by electron capture (EC) gas  chromatography.1

3.     INTERFERENCES

3.1     Method interferences may be caused by contaminants in solvents, reagents, glassware,
       and other sample-processing apparatus that lead to discrete artifacts or elevated baselines
       in gas chromatograms. All reagents and apparatus must be routinely demonstrated to
       be free from  interferences under the conditions of the analysis by running laboratory
       reagent blanks as described in Section 8.5.

       3.1.1  Glassware must be scrupulously cleaned.2  Clean all glassware as soon as possible
             after use by thoroughly rinsing with the last  solvent used  in it.  Follow by
             washing with hot water and detergent and thorough rinsing with dilute acid, tap
             and reagent water.  Drain dry, and heat in an oven or muffle furnace at 400°C for
             15 to 30 minutes. Do not heat volumetric ware.  Thermally stable materials, such
             as PCBs, may not be eliminated by this treatment. Thorough rinsing with acetone
             and pesticide-quality hexane may be substituted for the heating.  After  drying
             and cooling, seal  and store glassware in a clean environment to prevent any
             accumulation of dust or other  contaminants.   Store inverted or capped with
             aluminum foil.

       3.1.2  The use of high-purity reagents  and solvents  helps to minimize interference
             problems.   Purification  of solvents by distillation in all-glass systems may be
             required.

3.2     The acid forms of the herbicides are  strong organic acids, which react readily with
       alkaline  substances and can be lost during analysis.  Glassware and glass  wool must be
       acid-rinsed with (1+9) hydrochloric acid and the sodium sulfate must be acidified with
       sulfuric  acid prior to use to avoid this possibility.

3.3     Organic acids and phenols, especially chlorinated compounds, cause  the most direct
       interference with the determination.  Alkaline hydrolysis and subsequent extraction of
       the basic solution remove many chlorinated hydrocarbons and phthalate esters that might
       otherwise interfere with the electron capture analysis.

3.4     Matrix interferences may be caused by contaminants  that are coextracted  from the
       sample.  The extent of matrix interferences will vary considerably from source to  source,
       depending  upon the nature and  diversity of the industrial complex or municipality
       sampled. The cleanup procedure in Section 11 can be used to overcome many of these

-------
                                                                                Method 615
       interferences, but unique samples may require additional cleanup approaches to achieve
       the MDL listed in Table 1.

4.     SAFETY

4.1     The toxicity or carcinogenicity of each reagent used in this method has not been precisely
       defined; however, each chemical compound must be treated as a potential health hazard.
       From this viewpoint, exposure to these chemicals must be reduced to the lowest possible
       level by whatever means available.  The laboratory is responsible for maintaining a
       current awareness file  of OSHA regulations regarding the safe handling of the chemicals
       specified in this method. A reference file of material data handling sheets should also
       be made available  to all personnel involved  in the  chemical  analysis.   Additional
       references to  laboratory  safety are  available and have  been identified35  for  the
       information of the analyst.

4.2     Diazomethane is  a  toxic carcinogen and  can explode  under certain conditions.  The
       following precautions  must be followed:

       4.2.1  Use only a well-ventilated hood; do not breath vapors.

       4.2.2  Use a safety screen.

       4.2.3  Use mechanical pipetting aides.

       4.2.4  Do not heat above 90°C: EXPLOSION may result.

       4.2.5  Avoid grinding surfaces, and avoid the use of ground-glass joints, sleeve bearings,
             and glass stirrers: EXPLOSION may result.

       4.2.6  Do not store near alkali metals:  EXPLOSION may result.

       4.2.7  Solutions of diazomethane decompose rapidly in the presence of solid materials
             such as copper powder, calcium chloride, and boiling chips.

5.     APPARATUS AND MATERIALS

5.1     Sampling equipment,  for discrete or composite sampling.

       5.1.1  Grab-sample bottle:  Amber borosilicate or  flint glass, 1-L or 1-quart volume,
             fitted  with screw-caps lined with TFE-fluorocarbon.  Aluminum foil may  be
             substituted for TFE if the  sample is not corrosive.   If amber bottles are  not
             available, protect samples  from light.   The container and  cap liner must  be
             washed, rinsed with acetone or methylene  chloride, and dried before  use to
             minimize contamination.

       5.1.2  Automatic sampler (optional):  Must incorporate glass sample containers for the
             collection of a minimum of 250 mL. Sample containers must be kept refrigerated
             at 4°C and protected from light  during compositing.  If the sampler  uses a
             peristaltic pump, a minimum length of compressible silicone  rubber tubing may

-------
Method 615
              be used. Before use, however, the compressible tubing must be thoroughly rinsed
              with methanol, followed by repeated rinsings with reagent water to minimize the
              potential for contamination of the sample. An integrating flow meter is required
              to collect flow-proportional composites.

5.2     Glassware.   (All  specifications  are  suggested.   Catalog numbers  are  included for
       illustration  only.)

       5.2.1   Separatory  funnels:  60-mL  and 2000-mL,  with TFE-fluorocarbon stopcocks,
              ground- glass or TFE stoppers.

       5.2.2   Concentrator tube, Kuderna-Danish:  10-mL, graduated (Kontes K-570050-1025 or
              equivalent).  Calibration must be checked at the volumes employed in the test.
              Ground-glass stopper is used to prevent evaporation  of extracts.

       5.2.3   Evaporative  flask,   Kuderna-Danish:    500-mL  (Kontes  K-570001-0500  or
              equivalent).  Attach to concentrator tube with springs.

       5.2.4   Snyder column,  Kuderna-Danish:  Three-ball macro  (Kontes K-503000-0121 or
              equivalent).

       5.2.5   Snyder column,  Kuderna-Danish:   Two-ball micro  (Kontes  K-569001-0219 or
              equivalent).

       5.2.6   Erlenmeyer flask: Pyrex,  250-mL with 24/40 ground-glass joint.

       5.2.7   Vials: Amber glass, 10- to 15-mL capacity with TFE-fluorocarbon-lined screw-cap.

5.3     Boiling chips:  Approximately  10/40 mesh.  Heat at 400°C for 30 minutes or perform a
       Soxhlet extraction with methylene chloride.

5.4     Water bath: Heated, with concentric ring cover, capable of temperature control  (±2°C).
       The bath should be used in a hood.

5.5     Balance: Analytical,  capable of accurately weighing to the nearest 0.0001 g.

5.6     Diazomethane generator: assemble from two test tubes 150 mm  long by 20 mm ID, two
       Neoprene rubber stoppers, and a source of nitrogen.  The generator assembly is shown
       in Figure 1.

5.7     Glass wool: Acid-washed (Supelco 2-0383 or equivalent).

5.8     Gas chromatograph:  Analytical  system complete with gas chromatograph suitable for
       on-column  injection and all required accessories including  syringes, analytical columns,
       gases, detector, and strip-chart recorder.  A data system is recommended for measuring
       peak areas.

-------
                                                                               Method 615
       5.8.1  Column 1:  180 cm long by 4  mm ID glass, packed with 1.5% SP-2250/1.95%
             SP-2401 on Supelcoport (100/120 mesh) or equivalent. This column was used to
             develop the method performance statements in Section 16.  Alternative columns
             may be used in accordance with the provisions described in Section  13.1.

       5.8.2  Column 2:  180 cm long by 4 mm ID glass,  packed with 5% OV-210 on Gas
             Chrom Q (100/120 mesh) or equivalent.

       5.8.3  Column 3:   180  cm long by  2 mm ID glass,  packed with  0.1% SP-1000 on
             Carbopak C (80/100 mesh) or equivalent.

       5.8.4  Detector:  Electron capture. This detector has proven effective in the analysis of
             wastewaters for the parameters listed in the scope and was used to develop the
             method performance statements in Section 15.  Alternative detectors, including
             a mass spectrometer, may be used in accordance with the provisions described
             in Section 13.1.

6.     REAGENTS

6.1     Reagent water:  Reagent water  is defined as a water in which an  interferant is  not
       observed at the method detection limit of each parameter of interest.

6.2     Acetone, hexane, methanol:  Pesticide-quality or equivalent.

6.3     Ethyl ether:  Nanograde,  redistilled in glass if necessary.  Must be free of peroxides as
       indicated  by EM Quant test strips   (available  from Scientific Products Co.,  Cat.
       No. PI 126-8, and other suppliers). Procedures recommended for removal of peroxides
       are provided with the test strips.  After cleanup, 20 mL ethyl alcohol preservative must
       be added to  each liter of ether.

6.4     Sodium sulfate:  ACS, granular,  acidified, anhydrous.  Condition heating in a shallow
       tray at 400°C for a  minimum of 4  hours to  remove phthalates  and other  interfering
       organic substances.  Alternatively, heat 16 hours at 450 to 500°C in a shallow tray or
       perform a Soxhlet extraction with methylene chloride for 48 hours. Acidify by slurrying
       100  g sodium sulfate with enough ethyl ether to just cover the solid.  Add 0.1  mL
       concentrated sulfuric acid and mix thoroughly. Remove the ether under vacuum.  Mix
       1 g of the resulting solid with 5 mL of reagent water and measure the pH of the mixture.
       It must be below pH 4. Store at 130°C.

6.5     Hydrochloric acid (1+9):   Add  one volume  of concentrated acid (ACS) to  9 volumes
       reagent water.

6.6     Potassium hydroxide solution: 37% aqueous solution (w/v). Dissolve 37 g  ACS-grade
       potassium hydroxide pellets in reagent water and dilute to 100 mL.

6.7     Sulfuric acid solution (1 + 1): Slowly add 50 mL H2SO4 (sp. gr. 1.84) to 50 mL of reagent
       water.

-------
Method 615
6.8     Sulfuric acid solution (1+3):  Slowly add 25 mL H2SO4 (sp. gr. 1.84) to 75 mL of reagent
       water.  Maintain at 4°C.

6.9     Carbitol: Diethylene  glycol monoethyl ether, ACS. Available from Aldrich Chemical Co.

6.10    Diazald:  N-methyl-JV-nitroso-p-toluenesulfonamide, ACS.  Available  from  Aldrich
       Chemical Co.

6.11    Silicic acid: Chromatographic grade, nominal 100 mesh. Store  at 130°C.

6.12    Stock standard solutions (1.00 ug/uL): Stock standard solutions can be prepared from
       pure standard materials or purchased as certified solutions.

       6.12.1  Prepare stock standard solutions by accurately weighing about 0.0100 g of pure
              acids.  Dissolve the material in pesticide-quality ethyl ether and dilute to volume
              in a 10-mL volumetric flask. Larger volumes can be used at the convenience of
              the analyst. If compound purity is certified at 96% or greater, the weight can be
              used without correction to calculate the  concentration of the stock  standard.
              Commercially prepared stock standards can be used at any concentration if they
              are certified by the manufacturer or by an independent  source.

       6.12.2  Transfer the stock standard solutions into PTFE-sealed screw-cap vials.  Store at
              4°C and protect from light.   Stock  standard solutions  should  be checked
              frequently  for  signs of degradation or evaporation,  especially just prior to
              preparing calibration standards from them.

       6.12.3  Stock standard solutions must be replaced after 1 week,  or sooner if comparison
              with check standards indicates a problem.

7.     CALIBRATION

7.1     Establish gas chromatographic operating  parameters equivalent to those indicated in
       Table 1. The gas chromatographic system must be calibrated using the external standard
       technique.

7.2     External standard calibration procedure.

       7.2.1   For each parameter of interest, prepare working standards of the free acids at a
              minimum of three concentration levels by adding accurately measured volumes
              of one  or more stock standards to a 10-mL volumetric  flask containing 1.0 mL
              methanol and diluting to volume with ethyl ether. One of the external standards
              should be representative of a concentration near, but above, the method detection
              limit.  The other concentrations should correspond to the range of concentrations
              expected in the sample  concentrates  or should define the working range of the
              detector.

       7.2.2   Prepare calibration standards by esterification of 1.00-mL  volumes of the working
              standards as  described in Section 11.  Using injections of 2 to 5 uL  of each
              calibration standard, tabulate peak height or area responses against the  mass of

-------
                                                                                Method 615
              free acid represented by the injection.  The results can be used  to prepare  a
              calibration curve for each parameter.  Alternatively, the ratio of the response to
              the mass injected, defined as the calibration factor (CF), can be calculated for each
              parameter at each standard concentration.  If the relative standard deviation of
              the  calibration  factor is less than  10% over the working  range, the average
              calibration factor can be used in place of a calibration curve.

       7.2.3   The working calibration curve or calibration factor must be verified on each
              working shift by the preparation of one or more calibration standards.   If the
              response for any parameter varies  from the predicted response by more than
              ±10%, the test must be repeated using a fresh calibration standard. Alternatively,
              a new calibration curve or calibration factor must be prepared for that parameter.

7.3     Before using any cleanup procedure, the analyst must process a series of calibration
       standards through  the procedure  to validate elution patterns and the absence  of
       interference from the reagents.

8.     QUALITY CONTROL

8.1     Each  laboratory using this method is required  to operate a formal quality control
       program. The minimum requirements of this program consist of an initial demonstration
       of laboratory capability and the analysis of spiked samples as a continuing check on
       performance. The laboratory is required to maintain performance records to define the
       quality of data  that is generated.

       8.1.1   Before performing any  analyses, the analyst must demonstrate  the ability to
              generate acceptable accuracy and precision with this  method.  This ability is
              established as described in Section 8.2.

       8.1.2   In recognition of the rapid advances occurring in chromatography,  the analyst is
              permitted certain  options  to  improve  the separations  or lower the cost  of
              measurements.  Each time such modifications to the method are made, the analyst
              is required to repeat the procedure in Section 8.2.

       8.1.3   The laboratory must spike and analyze a  minimum of 10%  of all samples to
              monitor continuing  laboratory performance.  This  procedure is  described in
              Section 8.4.

8.2     To establish the ability to generate acceptable accuracy and precision, the analyst must
       perform the following operations.

       8.2.1   Select a  representative spike concentration for each compound (acid or ester) to
              be measured.  Using stock standards, prepare a quality control  check sample
              concentrate  in acetone, 1000 times  more concentrated  than  the  selected
              concentrations.

       8.2.2   Using a pipette, add 1.00  mL of the check sample concentrate  to each of a
              minimum of  four  1000-mL  aliquots  of  reagent water.    A representative
              wastewater may be used in place of the reagent water, but one or more additional

-------
Method 615
             aliquots must be analyzed to determine background levels, and the spike level
             must exceed twice the background level for the test to be valid.  Analyze  the
             aliquots according to the method beginning in Section 10.

       8.2.3  Calculate the average percent recovery (R), and the standard deviation of the
             percent recovery (s), for the results.  Wastewater background corrections must be
             made before R and s calculations are performed.

       8.2.4  Using the appropriate data from Table 2, determine the  recovery and single
             operator precision  expected for the method, and compare these results to  the
             values calculated in Section 8.2.3.  If the data are not comparable, review potential
             problem areas and repeat the test.

8.3     The analyst must calculate method performance criteria and define the performance of
       the laboratory for each spike concentration and parameter being measured.

       8.3.1  Calculate upper and lower control limits for method performance as follows:

                    Upper Control Limit (UCL)  = R + 3s
                    Lower Control Limit (LCL)  = R - 3s

             where R and s are calculated as in Section 8.2.3.  The UCL and LCL can be used
             to construct control charts6 that are useful in observing trends in performance.

       8.3.2  The laboratory  must develop  and maintain  separate accuracy  statements of
             laboratory performance for wastewater samples. An accuracy statement for the
             method is defined as R ± s.  The accuracy statement should be developed by the
             analysis of four  aliquots of wastewater as described in Section 8.2.2, followed by
             the calculation of R and s. Alternatively, the analyst may  use four wastewater
             data points gathered through the requirement for continuing quality  control in
             Section 8.4. The accuracy statements should be updated regularly.6

8.4     The laboratory is required to collect in duplicate a portion of their samples to monitor
       spike recoveries.  The frequency of spiked sample analysis must be at least  10% of all
       samples or one spiked  sample per month, whichever  is greater.   One aliquot of the
       sample must be spiked and analyzed as described in Section  8.2.  If the recovery for a
       particular parameter does not fall within the control limits for method performance, the
       results reported for that parameter in all samples processed as part of the same set must
       be qualified as described in Section 14.3. The laboratory should monitor the  frequency
       of data so qualified to ensure that it remains at or below 5%.

8.5     Before processing any samples, the analyst must demonstrate through the analysis of a
       1-L  aliquot of reagent water that all  glassware and reagent interferences are under
       control.  Each time a set  of samples is extracted or there is a change in reagents, a
       laboratory reagent blank  must  be processed  as  a  safeguard  against laboratory
       contamination.

8.6     It is recommended that the laboratory adopt additional quality assurance practices for
       use with this method. The specific practices that are most productive depend upon the
       needs of the laboratory  and the nature of the samples.  Field duplicates may be analyzed

-------
                                                                               Method 615
       to  monitor the precision of the sampling technique.  When doubt  exists  over the
       identification of a  peak on the chromatogram, confirmatory techniques  such as gas
       chromatography with a  dissimilar  column,  specific  element  detector,  or  mass
       spectrometer must be used.  Whenever possible, the laboratory should perform analysis
       of quality control materials and participate in relevant performance evaluation studies.

9.     SAMPLE COLLECTION, PRESERVATION, AND HANDLING

9.1     Grab samples must be collected in glass containers.  Conventional sampling practices7
       should be followed;  however, the  bottle must not  be prerinsed with sample before
       collection.   Composite samples should be collected  in refrigerated glass containers in
       accordance with the requirements of the program. Automatic sampling equipment must
       be  as free as possible of plastic and other potential sources of contamination.

9.2     The samples must be  iced or refrigerated at 4°C  from  the  time of  collection until
       extraction.

9.3     All samples must be extracted within 7 days and completely analyzed within 40 days of
       extraction.

10.    SAMPLE EXTRACTION

10.1    Mark  the water  meniscus on the side of the sample bottle for later determination of
       sample volume. Pour the entire sample into a 2-L separatory funnel. Check the pH with
       wide-range pH paper and adjust to pH less than 2 with sulfuric acid (1 + 1).

10.2    Add 150 mL ethyl  ether to the sample bottle, cap the bottle, and shake 30 seconds to
       rinse the walls.  Transfer the solvent to the separatory funnel and extract the sample by
       shaking the funnel for 2 minutes with periodic venting to release excess pressure. Allow
       the organic layer to separate from the water phase for a minimum of 10 minutes. If the
       emulsion interface between layers is more than one-third the volume of the solvent layer,
       the analyst must employ mechanical techniques to complete the phase separation.  The
       optimum technique depends upon the sample, but may include stirring, filtration of the
       emulsion through glass wool, centrifugation, or other physical means. Drain the aqueous
       phase into a 1000-mL Erlenmeyer flask and collect the extract in a 250-mL ground-glass
       Erlenmeyer flask containing 2 mL of 37% potassium hydroxide solution.  Approximately
       80 mL of the ethyl ether will remain dissolved  in the aqueous  phase.

10.3    Add a 50-mL volume of ethyl ether to the sample  bottle  and repeat the extraction a
       second time, combining the extracts in  the Erlenmeyer flask. Perform a third extraction
       in the same manner.

10.4    Add 15 mL reagent water and one or  two clean boiling chips to the 250-mL flask and
       attach a three-ball Snyder column.  Prewet the Snyder column by adding 1 mL  ethyl
       ether to the top. Place the apparatus  on a hot water bath (60 to 65°C), such that the
       bottom of the flask is bathed in the water vapor. Although the ethyl ether will evaporate
       in about 15 minutes, continue heating for a total of 60 minutes, beginning from the time
       the flask is placed  on the water bath.  Remove the apparatus and let stand  at room
       temperature for at least 10 minutes.

-------
Method 615
10.5    Transfer the solution to a 60-mL separatory funnel using 5 to 10 mL of reagent water.
       Wash the basic solution twice by shaking for one minute with 20-mL portions of ethyl
       ether. Discard the organic phase. The free acids remain in the aqueous phase.

10.6    Acidify the contents of the separatory funnel to pH 2 by adding 2 mL of cold (4°C)
       sulfuric acid (1+3).  Test with pH indicator paper. Add 20 mL ethyl ether and shake
       vigorously for 2  minutes.  Drain the aqueous layer into the 250-mL Erlenmeyer flask,
       then pour the  organic layer into a 125-mL Erlenmeyer flask containing about 0.5 g of
       acidified  anhydrous sodium sulfate.  Repeat the  extraction  twice more with  10-mL
       aliquots of ethyl  ether, combining all solvent in the 125-mL flask.  Allow the extract to
       remain in contact with the sodium sulfate for approximately 2 hours.

10.7    Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10-mL concentrator tube
       to a 500- mL evaporative flask.  Other concentration devices or techniques may be used
       in place of the K-D if the requirements of Section 8.2 are met.

10.8    Pour the combined extract through a funnel plugged with acid-washed glass wool, and
       collect the extract in the K-D in concentrator. Use a glass rod to crush any caked sodium
       sulfate during the transfer.  Rinse the Erlenmeyer flask and column with 20 to 30 mL of
       ethyl ether to complete the quantitative transfer.

10.9    Add one to two clean boiling chips to the evaporative flask and attach a three-ball Snyder
       column. Prewet  the Snyder column  by adding about 1-mL ethyl ether to the top.  Place
       the  K-D apparatus  on a hot water  bath, 60 to 65°C, so that the concentrator tube is
       partially immersed in the hot water,  and  the entire lower rounded surface of the flask is
       bathed with hot vapor.  Adjust the vertical position of the  apparatus and the water
       temperature as required to complete the concentration in 15 to 20 minutes.  At the proper
       rate of distillation the balls of the column will actively chatter but the chambers will not
       flood. When the  apparent volume of liquid reaches 1 mL, remove the K-D apparatus and
       allow it to drain  and cool for at least 10  minutes.

10.10   Remove the Snyder column and rinse the flask and its lower joint into the concentrator
       tube with 1 to  2 mL of ethyl ether.  A 5-mL syringe is recommended for this operation.
       Add a fresh boiling chip.  Attach a micro-Snyder column to the concentrator tube and
       prewet the column by adding about 0.5  mL of ethyl ether to  the top.  Place the micro
       K-D apparatus on the water bath so that the concentrator tube is partially immersed in
       the hot water.  Adjust the vertical position of the apparatus and the water temperature
       as required to  complete concentration in 5 to 10 minutes. When the apparent volume
       of liquid reaches  0.5 mL, remove the micro K-D from the bath  and allow it to drain and
       cool.  Remove the micro Snyder column and add 0.1 mL of methanol.  Rinse the walls
       of the concentrator tube while adjusting  the volume to  1.0 mL with ethyl ether.

11.    ESTERIFICATION OF ACIDS

11.1    Assemble the diazomethane generator (see Figure 1) in a hood using two test tubes 150
       mm long by 20 mm ID.  Use neoprene  rubber stoppers with holes drilled in them to
       accommodate glass delivery tubes.  The exit tube must be drawn to a point to bubble
       diazomethane through the sample extract.

-------
                                                                              Method 615
11.2   Add 5 mL of ethyl ether to the first test tube. Add 1 mL of ethyl ether, 1 mL of carbitol,
      1.5 mL of 37% aqueous KOH, and 0.1-0.2 g Diazald to the second test tube. Immediately
      place the exit tube into the concentrator  tube containing the sample extract. Apply
      nitrogen flow (10 mL/min.) to bubble diazomethane through the extract for 10 minutes
      or until the yellow color of diazomethane  persists.

11.3   Remove the concentrator  tube and seal it with a neoprene or PTFE stopper.  Store at
      room temperature in a hood for 20 minutes.

11.4   Destroy any  unreacted diazomethane by  adding  0.1 to 0.2  g  silicic acid to  the
      concentrator tube.  Allow to stand until  the evolution of nitrogen gas has  stopped.
      Adjust the sample volume to 10.0 mL with hexane.  Stopper the concentrator tube and
      store refrigerated if further processing will  not  be performed immediately.  It is
      recommended that the methylated extracts be analyzed immediately to  minimize any
      transesterification and  other potential reactions that may occur.   Analyze by  gas
      chromatography.

11.5   Determine the original sample  volume by refilling the sample bottle to  the mark and
      transferring the water to a 1000-mL graduated cylinder. Record the sample volume to
      the nearest 5 mL.

12.   CLEANUP AND SEPARATION

12.1   No cleanup procedures were required to analyze the wastewaters described in Section 16.
      If particular circumstances demand the use of a cleanup procedure, the analyst must
      determine the elution profile and demonstrate that the recovery of each compound of
      interest for the cleanup procedure is no less than 85%.

13.   GAS CHROMATOGRAPHY

13.1   Table 1 summarizes the recommended operating conditions for the gas chromatograph.
      Included in this table are estimated retention times and method detection  limits that  can
      be achieved by this method.  Examples of the separations achieved for the methyl esters
      are shown in Figures 2 to 3.  Other packed columns, chromatographic  conditions, or
      detectors may be  used  if  the requirements  of  Section 8.2  are  met.    Capillary
      (open-tubular) columns may also be used if the relative standard deviations of responses
      for replicate injections are demonstrated  to be less than 6%  and the requirements of
      Section 8.2 are met.

13.2   Calibrate the system daily as described in  Section 7.

13.3   Inject  1 to 5  uL of the sample  extract using the solvent-flush technique.8   Record  the
      volume injected to the nearest 0.05 uL, and the resulting peak size in area or peak height
      units.  An automated system that consistently injects a constant volume of extract may
      also be used.

13.4   The width of the retention-time window used to make identifications should  be based
      upon measurements of actual retention-time variations of standards over the course of
      a day.  Three times the standard deviation of a retention time can be used to  calculate

-------
Method 615
       a suggested window size for a compound.  However, the experience of the analyst
       should weigh heavily in the interpretation of chromatograms.

13.5    If the response for the peak exceeds the working range of the system, dilute the extract
       and reanalyze.

13.6    If the measurement of the peak response is prevented by the presence of interferences,
       further cleanup is required.

14.    CALCULATIONS

14.1    Determine the concentration of individual compounds in the  sample.   Calculate the
       amount of free acid injected from the peak response using the calibration curve or
       calibration factor in Section 7.2.2. The concentration in the sample can be calculated as
       follows:
                                       Equation 1

                                                    (A}
                             Concentration, \ig/L -
              where
              A  = Amount of material injected, in ng
              Vt  = Volume of extract injected, in uL
              Vt  = Volume of total extract, in uL
              V,  = Volume of water extracted, in ml
                                                   (V) (V)
14.2    Report results in micrograms per liter as acid equivalent without correction for recovery
       data. When duplicate and spiked samples are analyzed, report all data obtained with the
       sample results.

14.3    For samples processed as part of a set where the laboratory spiked sample recovery falls
       outside  of the control limits in Section 8.3, data for the affected parameters must be
       labeled as suspect.

-------
                                                                              Method 615
15.    GC/MS CONFIRMATION

15.1    It is recommended that GC/MS  techniques be judiciously  employed to support
       qualitative compound  identifications made with this method.  The mass spectrometer
       should be capable of scanning the mass range from 35 amu to a mass 50 amu above the
       molecular weight of the methyl ester of the acid herbicide.  The instrument must be
       capable of scanning the mass range at a rate to produce at least 5 scans per peak but not
       to exceed 7 seconds per scan utilizing a 70 V (nominal) electron energy in the electron
       impact ionization mode.  A GC-to-MS interface constructed of all glass or glass-lined
       materials  is  recommended.  A computer system should  be  interfaced to the mass
       spectrometer that allows the continuous acquisition  and storage on machine-readable
       media of all mass  spectra obtained throughout the  duration of the chromatographic
       program.

15.2    Gas chromatographic columns and conditions should be selected for optimum separation
       and performance. The conditions selected must be compatible with standard GC/MS
       operating practices. Chromatographic tailing factors of less  than 5.0 must be achieved.9

15.3    At the beginning of each day that confirmatory analyses are to be performed, the GC/MS
       system must be checked to  see  that  all decafluorotriphenyl  phosphine  (DFTPP)
       performance criteria are achieved.10

15.4    To confirm an identification of a  compound, the background-corrected mass spectrum
       of the methyl ester must be obtained from the sample extract and compared with  a mass
       spectrum from a stock or calibration  standard analyzed under the same chromatographic
       conditions.  It is recommended that at least 25 ng of material  be injected into the GC/MS.
       The criteria below must be met for qualitative confirmation.

       15.4.1 All ions that are present above 10% relative abundance in the mass spectrum of
             the standard must be present in the mass spectrum of the sample with agreement
             to  ±10%.  For example,  if the relative abundance of an ion is 30% in the mass
             spectrum of the standard,  the allowable limits for the relative abundance  of that
             ion in the mass spectrum for the sample would be 20 to 40%.

       15.4.2 The retention time of the compound  in the sample must be within 6 seconds of
             the same compound in the standard solution.

       15.4.3 Compounds that have very similar mass spectra can be explicitly identified by
             GC/MS only on the basis  of retention time data.

15.5    Where available, chemical ionization mass spectra  may be employed to aid  in  the
       qualitative identification process.

15.6    Should these MS procedures fail to provide satisfactory results, additional steps may be
       taken before reanalysis. These may include the use of alternate packed or capillary GC
       columns or additional  cleanup.

-------
Method 615
16.    METHOD PERFORMANCE

16.1    The method detection  limit (MDL) is  defined  as  the  minimum  concentration of a
       substance that can be measured and reported with 99% confidence that the value is above
       zero.11 The MDL concentrations listed in Table 1 were obtained from reagent water with
       an electron capture detector.1

16.2    In  a single laboratory (West Coast Technical Services, Inc.), using reagent water and
       effluents from publicly owned treatment works (POTW), the average recoveries presented
       in Table 2 were obtained.1 The standard deviations of the percent recoveries of these
       measurements are also included in Table 2.

-------
                                                                               Method 615
                                    References

1.      "Pesticide Methods Evaluation," Letter Report #33 for EPA Contract No. 68-03-2697.
       Available from U.S. Environmental Protection Agency, Environmental Monitoring and
       Support Laboratory, Cincinnati, Ohio.

2.      ASTM Annual Book of Standards,  Part 31, D3694, "Standard Practice for Preparation of
       Sample Containers and for Preservation," American Society for Testing and Materials,
       Philadelphia, PA, p.  679, 1980.

3.      "Carcinogens—Working  with  Carcinogens," Department of Health, Education, and
       Welfare, Public Health  Service,  Center for Disease Control,  National Institute for
       Occupational Safety and Health, Publication No. 77-206, August 1977.

4.      "OSHA Safety and Health Standards,  General Industry" (29 CFR 1910), Occupational
       Safety and Health Administration, OSHA 2206 (Revised, January  1976).

5.      "Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
       Committee on Chemical  Safety, 3rd Edition, 1979.

6.      "Handbook for  Analytical Quality Control in Water and Wastewater Laboratories,"
       EPA-600/4-79-019, U. S.  Environmental Protection Agency, Environmental Monitoring
       and Support Laboratory  - Cincinnati, Ohio 45268, March 1979.

7.      ASTM  Annual Book of Standards, Part 31, D3370, "Standard Practice for Sampling
       Water," American Society for Testing and Materials, Philadelphia, PA, p. 76,  1980.

8.      Burke,  J. A.  "Gas Chromatography for Pesticide Residue Analysis;  Some Practical
       Aspects," Journal of the Association of Official Analytical Chemists, 48, 1037  (1965).

9.      McNair, H.M. and  Bonelli,  E. J.  "Basic Chromatography,"  Consolidated Printing,
       Berkeley, California, p. 52, 1969.

10.     Eichelberger, J.W., Harris, L.E.,  and Budde, W.L. "Reference Compound to Calibrate Ion
       Abundance Measurement in  Gas Chromatography-Mass  Spectrometry,"  Analytical
       Chemistry, 47,  995 (1975).

11.     Glaser, J.A. et al. "Trace  Analysis for Wastewaters," Environmental Science & Technology,
       15, 1426 (1981).

-------
Method 615
Table 1.
Chromatographic Conditions and Method Detection  Limits
retention lime Method Detection
Column 1 Colu
1.2 1
2.0 1
mn 2 Column 3 Limit fyig/L)
.0 _ 0.27
.6 - 1.20
2.7 2.0 _ 0.17
3.4 2.4 _ 0.20
4.1

3.4
4.1
4.8
11.2
0.91
5.0 5.80
192.00
249.00
0.65
0.07
Parameter
(as methyl ester)
Dicamba
2,4-D
2,4,5-TP
2,4,5-T
2,4-DB
Dalapon
MCPP
MCPA
Dichlorprop
Dinoseb
Column 1 conditions: Supelcoport (100 / 120 mesh) coated with 1.5% SP-2250 / 1.95%  SP-2401
packed in a glass column 1.8 m long by 4 mm ID with 95% argon / 5% methane carrier  gas at a
flow rate of 70 mL / min. Column temperature: isothermal at 185°C, except for MCPP, MCPA,
dichlorprop and dinoseb, where the column temperature was held at 140°C for 6 minutes and
then programmed to 200°C at 10° / min.  An electron capture detector was used to measure MDL.

Column 2 conditions: Gas Chrom Q (100 / 120 mesh) coated with 5% OV-210 packed in a glass
column 1.8 m long by 4 mm ID  with 95% argon/5% methane carrier gas at a flow rate of
70 mL / min.  Column temperature: isothermal at 185°C.

Column 3 conditions: Carbopak C (80 /  100 mesh) coated with 0.1% SP-1000 packed in a glass
column 1.8 m long by 2 mm ID with nitrogen carrier gas at a flow rate of 25 mL / min.  Column
temperature:  programmed at injection from 100 to 150°C at 10°/min.

-------
                                                                                Method 615
Table 2. Single-Operator Accuracy and Precision*


Parameter
2,4-D


Dalapon


2,4-DB


Dicamba


Dichlorprop


Dinoseb

MCPA


MCPP


2,4,5-T


2,4,5-TP



Sample
Type
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW

Spike
(lig/L)
10.9
10.1
200.0
23.4
23.4
468.0
10.3
10.4
208.0
1.2
1.1
22.2
10.7
10.7
213.0
0.5
102.0
2020.0
2020.0
21400.0
2080.0
2100.0
20440.0
1.1
1.3
25.5
1.0
1.3
25.0

Mean Recovery
(%)
75
77
65
66
96
81
93
93
77
79
86
82
97
72
100
86
81
98
73
97
94
97
95
85
83
78
88
88
72
Standard
Deviation
(%)
4
4
5
8
13
9
3
3
6
7
9
6
2
3
2
4
3
4
3
2
4
3
2
6
4
5
5
4
5
*A11 results based upon seven replicate analyses.

DW = Reagent water
MW = Municipal water

-------
Method 615
                                                                         Glass Tubing
         Nitrogen
       Rubber Stopper
—
(•• **•








\J
¥
-N/>-
0
0
o
0 •
O
0 o
0
0 o
0
D
                                                          TO
                               Tubel
Tube 2
                             Figure 1.  Diazomethane Generator
                                                                                  A52-002-17A

-------
                                                                           Method 615
T
 0
                2,4-D
2.0
4.0
6.0
8.0
                                                                  10.0
                             Retention Time (minutes)
                                                                          A52-002-I6A
 Figure 2.   Gas Chromatogram of Methyl Esters of Chlorinated Herbicides on
             Column 1 (for conditions, see Table 1)

-------
Method 615
                            ,MCPA
                                                                  Dinoseb
               2.0
4.0
6.0
8.0
10.0
12.0
14.0
                                    Retention Time (minutes)
16.0
                                                                                  AS2-002-1SA
           Figure 3.  Gas Chromatogram of Methyl Esters of Chlorinated Herbicides
                     on Column 1 (for conditions, see Table 1)

-------