SEPA
United States
Environmental Protection
Agency
 www.epa.gov
   Method 619: The Determination of
   Triazine Pesticides in Municipal
   and Industrial Wastewater

-------
              Method 619
The Determination of Triazine
  Pesticides in Municipal and
       Industrial Wastewater

-------
                                 Method 619
        The Determination of Triazine Pesticides in Municipal and
                             Industrial Wastewater


1.     SCOPE AND APPLICATION

1.1     This method covers the  determination of certain triazine pesticides.  The following
       parameters can be determined by this  method:

             Parameter              STORETNo.            CAS No.
             Ametryn                       _              834-12-8
             Atraton                        -             1610-17-9
             Atrazine                    39033             1912-24-9
             Prometon                  39056             1610-18-0
             Prometryn                  39057             7287-19-6
             Propazine                  39024              139-40-2
             Secbumeton                    -            26259-45-0
             Simetryn                   39054             1014-70-6
             Simazine                   39055              122-34-9
             Terbuthylazine                  -             5915-41-3
             Terbutryn                      -               86-50-0
1.2    This is a gas  chromatographic (GC) method applicable  to the determination  of the
      compounds listed above in  industrial and  municipal discharges as provided  under
      40 CFR 136.1. Any modification of this method beyond those expressly permitted shall
      be considered a major modification subject to application and approval of alternative test
      procedures under 40 CFR 136.4 and 136.5.

1.3    The estimated method detection limit (MDL, defined in Section 15) for each parameter
      is listed in Table 1. The MDL for a specific wastewater may  differ from those listed,
      depending upon the nature of interferences in the sample matrix.

1.4    The sample extraction and concentration steps in this method are essentially the same as
      several others in 600-series methods. Thus, a single sample  may be extracted to measure
      the parameters included in the  scope  of each of these  methods. When cleanup  is
      required, the concentration  levels must be high enough to permit selecting aliquots, as
      necessary,  in   order  to  apply  appropriate   cleanup  procedures.  Under  gas
      chromatography, the analyst is allowed the latitude to select chromatographic conditions
      appropriate for the simultaneous measurement of combinations of these parameters (see
      Section 12).

1.5    This method is restricted to use by or under the supervision of analysts experienced in
      the use of gas chromatography and in the interpretation of gas chromatograms. Each
      analyst must demonstrate  the ability to generate  acceptable results with this method
      using the procedure described in Section 8.2.

-------
Method 619
1.6     When this method is used to analyze unfamiliar samples for any or all of the compounds
       above, compound identifications should be  supported by  at  least one additional
       qualitative technique.  This  method describes analytical conditions for a second gas
       chromatographic column that can be used to confirm measurements made with the
       primary column. Section 14 provides gas chromatograph/mass spectrometer (GC/MS)
       criteria appropriate for the qualitative confirmation of compound identifications.

2.     SUMMARY OF METHOD

2.1     A measured volume of sample,  approximately 1  L, is extracted with 15% methylene
       chloride using  a  separatory  funnel.   The methylene chloride  extract is dried  and
       exchanged  to  hexane  during concentration  to  a volume of 10  mL or less.   Gas
       chromatographic conditions are described which permit the separation and measurement
       of the compounds in the extract by gas chromatography with a thermionic bead detector
       in the nitrogen mode.1'2

2.2     Method 619 represents  an editorial revision  of  a previously  promulgated U.S. EPA
       method for organophosphorus pesticides.3 While complete method validation data is not
       presented herein, the method has been in widespread use since its promulgation,  and
       represents the state of the art for the analysis of such materials.

2.3     This  method provides an optional  Florisil column cleanup procedure to aid  in the
       elimination or reduction of interferences which may be encountered.

3.     INTERFERENCES

3.1     Method interferences may be caused by contaminants in solvents, reagents, glassware,
       and other sample-processing apparatus that lead to discrete artifacts or elevated baselines
       in gas chromatograms. All reagents and apparatus must be routinely demonstrated to
       be  free from interferences under the conditions of the  analysis by running laboratory
       reagent blanks as described in Section 8.5.

       3.1.1    Glassware must be scrupulously cleaned.4 Clean all glassware as soon as possible
              after  use by thoroughly rinsing with the last solvent used in it.   Follow by
              washing with hot water and detergent and thorough rinsing with tap and reagent
              water.  Drain  dry, and heat in an oven or muffle furnace at 400°C  for 15 to 30
              minutes. Do not heat volumetric ware.  Thermally stable materials, such as PCBs,
              may not be eliminated by this treatment.  Thorough rinsing with acetone  and
              pesticide-quality hexane may be substituted for the heating. After drying  and
              cooling,  seal  and  store glassware  in  a   clean environment to prevent  any
              accumulation  of dust  or  other contaminants.  Store inverted or capped with
              aluminum  foil.

       3.1.2    The use of high-purity reagents and solvents helps to minimize  interference
              problems.  Purification of solvents by distillation in all-glass systems may be
              required.

3.2     Matrix interferences  may  be caused by  contaminants that  are coextracted from the
       sample. The extent of matrix interferences will vary considerably from source to source,

-------
                                                                               Method 619
       depending  upon the nature and diversity of the  industrial complex or municipality
       sampled. The cleanup procedure in Section 11 can be used to overcome many of these
       interferences, but unique samples may require additional cleanup approaches to achieve
       the MDL listed in Table 1.

4.     SAFETY

4.1     The toxicity or carcinogenicity of each reagent used in this method has not been precisely
       defined; however, each chemical compound must be treated as a potential health hazard.
       From this viewpoint, exposure to these chemicals must be reduced to the lowest possible
       level by whatever  means available.  The laboratory  is responsible for maintaining a
       current awareness file of OSHA regulations regarding the safe handling of the chemicals
       specified in this method. A reference file of material data handling sheets should also
       be made available  to  all personnel involved  in  the  chemical  analysis.   Additional
       references to  laboratory  safety  are  available and  have  been identified57  for  the
       information of the analyst.

5.     APPARATUS AND MATERIALS

5.1     Sampling equipment, for discrete or composite sampling.

       5.1.1  Grab-sample bottle:  Amber borosilicate or flint glass, 1-L or 1-quart volume,
             fitted with screw-caps  lined with TFE-fluorocarbon.  Aluminum foil  may  be
             substituted  for TFE if  the sample is  not corrosive.  If amber  bottles are  not
             available, protect samples from light.  The container  and  cap liner must  be
             washed, rinsed  with acetone or methylene chloride, and dried before use to
             minimize contamination.

       5.1.2  Automatic sampler (optional):  Must incorporate glass sample containers for the
             collection of a minimum of 250 mL. Sample containers must be kept refrigerated
             at 4°C  and protected from light during compositing.   If the sampler uses a
             peristaltic pump, a minimum length of compressible silicone  rubber tubing may
             be used. Before use, however, the compressible tubing must be thoroughly rinsed
             with methanol, followed by repeated rinsings with reagent water  to minimize the
             potential for  contamination of the sample. An integrating  flow meter is required
             to collect flow-proportional composites.

5.2     Glassware.  (All  specifications  are  suggested.  Catalog  numbers are  included  for
       illustration only.)

       5.2.1  Separatory funnel:  2000-mL, with TFE-fluorocarbon stopcock, ground-glass or
             TFE stopper.

       5.2.2  Drying column:   Chromatographic column 400 mm long by 19 mm ID with
             coarse-fritted disc.

       5.2.3  Chromatographic  column: 400 mm long by 19 mm ID with coarse-fritted disc at
             bottom and TFE-fluorocarbon stopcock (Kontes K-420540-0224 or equivalent).

-------
Method 619
       5.2.4  Concentrator tube, Kuderna-Danish:  10-mL, graduated (Kontes K-570050-1025 or
             equivalent).  Calibration must be checked at the volumes employed in the test.
             Ground-glass stopper is used to prevent evaporation of extracts.

       5.2.5  Evaporative   flask,  Kuderna-Danish:    500-mL  (Kontes  K-570001-0500  or
             equivalent).  Attach to concentrator tube with springs.

       5.2.6  Snyder column, Kuderna-Danish: Three-ball macro  (Kontes K-503000-0121 or
             equivalent).

       5.2.7  Vials:  Amber glass, 10- to 15-mL capacity with TFE-fluorocarbon-lined screw-cap.

5.3     Boiling chips:  Approximately  10/40 mesh.  Heat at 400°C for 30 minutes or perform a
       Soxhlet extraction with methylene chloride.

5.4     Water bath:  Heated, with concentric ring cover, capable of temperature control (±2°C).
       The bath should be used in a hood.

5.5     Balance:  Analytical,  capable of accurately weighing to the nearest 0.0001 g.

5.6     Gas chromatograph:  Analytical system complete with gas chromatograph suitable for
       on-column injection and all required accessories including syringes, analytical columns,
       gases, detector, and strip-chart  recorder.  A data system is recommended for measuring
       peak areas.

       5.6.1  Column 1: 180 cm long by 2 mm ID glass, packed with  5% Carbowax 20M-TPA
             on Supelcoport (80/100 mesh) or equivalent.  This column was used to develop
             the method performance statements in Section 15.  Alternative columns may be
             used in accordance with the provisions described in Section 12.1.

       5.6.2  Column 2: 180 cm long by 4 mm ID glass, packed with 1.0% Carbowax 20M on
             Gas Chrom Q (100/120  mesh) or equivalent.

       5.6.3  Detector:  Thermionic bead in the nitrogen  mode.   This detector has proven
             effective in the analysis  of wastewaters for the parameters listed in the scope and
             was used to develop  the method  performance  statements  in Section  15.
             Alternative detectors, including a mass spectrometer, may be used in accordance
             with the provisions described in Section 12.1.

6.     REAGENTS

6.1     Reagent water:  Reagent water is  defined  as a water  in which an interferant is not
       observed at the method detection limit of each parameter of interest.

6.2     Acetone, hexane, methylene chloride, methanol: Pesticide-quality or equivalent.

6.3     Ethyl ether:  Nanograde, redistilled in glass if necessary. Must be free of peroxides as
       indicated  by  EM  Quant test strips  (available  from Scientific  Products  Co.,  Cat.
       No. PI 126-8, and other suppliers).  Procedures recommended for removal of peroxides

-------
                                                                                Method 619
       are provided with the test strips.  After cleanup, 20 mL ethyl alcohol preservative must
       be added to each liter of ether.

6.4     Sodium sulfate:  ACS, granular, anhydrous. Condition by heating in a shallow tray at
       400°C for  a minimum of 4 hours to remove phthalates and other interfering organic
       substances.  Alternatively, heat 16 hours at 450 to 500°C in a shallow tray or perform a
       Soxhlet extraction with methylene chloride for 48 hours.

6.5     Florisil: PR grade (60/100 mesh).  Purchase activated at 675°C and store in the dark in
       glass container with ground-glass stopper or foil-lined screw-cap. Before use, activate
       each batch at least 16 hours at 130°C in a foil-covered glass container.

6.6     Stock standard solutions  (1.00 ug/uL):  Stock standard solutions may be prepared from
       pure standard materials or purchased as certified solutions.

       6.6.1   Prepare stock standard solutions by accurately weighing approximately 0.0100 g
              of  pure material.  Dissolve the material in pesticide-quality hexane  or  other
              suitable solvent and dilute to  volume in a 10-mL volumetric flask.  Larger
              volumes may be used at the convenience of the analyst.  If compound purity is
              certified at 96% or greater,  the weight may be used without correction to calculate
              the concentration of the stock standard.  Commercially prepared stock standards
              may be used at any concentration if they are certified by the manufacturer or by
              an  independent source.

       6.6.2   Transfer the stock standard solutions  into  TFE-fluorocarbon-sealed screw-cap
              vials.  Store at 4°C and protect from  light.  Frequently  check stock  standard
              solutions  for signs of degradation or  evaporation, especially  just  prior  to
              preparing calibration standards from them.

       6.6.3   Stock standard solutions must be replaced after 6 months, or sooner if comparison
              with check standards indicates a problem.

7.     CALIBRATION

7.1     Establish gas chromatographic operating parameters equivalent to  those indicated in
       Table 1. The gas chromatographic system may be calibrated using  either the external
       standard technique  (Section 7.2) or the internal standard technique (Section 7.3).

7.2     External standard calibration procedure

       7.2.1   For each parameter of  interest, prepare calibration standards at a minimum of
              three concentration levels by adding accurately measured volumes of one or more
              stock standards to a volumetric flask and diluting to volume with hexane or other
              suitable solvent.  One  of  the external standards should be representative of a
              concentration  near,  but  above,  the  method  detection limit.    The  other
              concentrations should correspond to the range of concentrations expected in the
              sample concentrates or should define the working  range of the detector.

-------
Method 619
       7.2.2   Using injections of 1 to 5 uL of each calibration standard, tabulate peak height or
              area responses against the mass injected.  The results can be used to prepare a
              calibration curve for each parameter.  Alternatively, the ratio of the response to
              the mass injected,  defined as the calibration factor (CF), may be calculated for
              each parameter at each standard concentration. If the relative standard deviation
              of the calibration factor is less than 10% over the working range, the average
              calibration factor can be used in place of a calibration curve.

       7.2.3   The working calibration curve  or calibration factor must be verified on each
              working shift by the measurement of one or more calibration standards.  If the
              response for any parameter varies  from the predicted response by more than
              ±10%, the  test must be repeated using a fresh calibration standard.  Alternatively,
              a new calibration curve or calibration factor must be prepared for that parameter.

7.3     Internal standard calibration procedure: To use this approach,  the analyst must select
       one or more internal standards similar in analytical behavior to the compounds of
       interest.  The analyst must  further demonstrate that the measurement of the internal
       standard is not affected by method or matrix interferences.  Due to these limitations, no
       internal standard applicable to all samples can be suggested.

       7.3.1   Prepare calibration standards at a minimum of three concentration levels for each
              parameter of interest by adding volumes of one or more stock standards  to a
              volumetric flask. To each calibration standard, add a known constant amount of
              one or more internal standards, and  dilute  to volume with hexane or other
              suitable solvent.  One of the standards should be representative of a concentration
              near, but above, the method detection  limit.  The other concentrations should
              correspond to the range of concentrations expected in the sample concentrates, or
              should define the working range of the detector.

       7.3.2   Using injections of 1 to  5 uL of each calibration standard, tabulate the peak height
              or area responses against the  concentration  for each compound and internal
              standard.  Calculate response factors (RF) for each compound as follows:


                                        Equation 1

                                            (A ) (C.)
                                     RF =  V s} ^ ls'
              where
              As = Response for the parameter to be measured
              Als = Response for the internal standard
              Cis = Concentration of the internal standard, in ug/L
              Cs = Concentration of the parameter to be measured, in ug/L

-------
                                                                                Method 619
              If the RF value over the working range  is constant,  less than 10% relative
              standard deviation, the RF can be assumed to be invariant and the  average RF
              may  be used for calculations.  Alternatively,  the results may be used to plot a
              calibration curve of response ratios, As/Ais against RF.

       7.3.3   The working calibration curve or RF must be verified on each working shift by
              the measurement of one or more calibration standards.  If the response for any
              parameter varies from the predicted response by more than ±10%, the test  must
              be repeated using a fresh calibration standard.  Alternatively, a new calibration
              curve must be prepared for that  compound.

7.4     The cleanup procedure in Section 11 utilizes Florisil  chromatography.  Florisil  from
       different batches or sources may vary in adsorptive capacity. To standardize the amount
       of Florisil which is used, the use of the lauric acid value is suggested.  This procedure8
       determines the adsorption from hexane  solution of lauric acid, in milligrams, per  gram
       of Florisil. The amount of Florisil to  be  used for each column is calculated by dividing
       this factor into 110 and multiplying by 20 g.

7.5     Before using any cleanup procedure, the  analyst must process a  series of calibration
       standards through  the procedure to validate  elution patterns and  the  absence of
       interference  from the reagents.

8.     QUALITY CONTROL

8.1     Each  laboratory using this  method  is  required to  operate  a formal quality control
       program. The minimum requirements of this program consist of an initial demonstration
       of laboratory capability and the analysis of spiked samples as a continuing check on
       performance. The laboratory is required to maintain performance records to define the
       quality of data that is generated.

       8.1.1   Before performing any analyses, the analyst must demonstrate  the  ability to
              generate  acceptable accuracy and precision with this method.  This ability is
              established as described in Section  8.2.

       8.1.2   In recognition of the rapid advances occurring in chromatography, the analyst is
              permitted certain options to  improve  the  separations  or lower the cost of
              measurements. Each time such modifications to the method are made, the analyst
              is required to repeat the procedure in Section 8.2.

       8.1.3   The laboratory must spike and  analyze  a minimum of 10% of all samples to
              monitor continuing laboratory performance.  This procedure  is described in
              Section 8.4.

8.2     To establish  the ability to generate acceptable accuracy and precision, the analyst  must
       perform the  following operations.

       8.2.1   Select a representative spike concentration for each compound to be measured.
              Using stock standards, prepare  a quality control check sample concentrate in
              acetone, 1000 times more concentrated than the selected concentrations.

-------
Method 619
       8.2.2  Using a pipette, add  1.00 mL of the check sample concentrate to each  of a
             minimum  of four  1000-mL  aliquots  of reagent  water.   A representative
             wastewater may be used in place of the reagent water, but one or more additional
             aliquots must be analyzed to determine background levels, and the spike level
             must exceed twice the  background level  for the test to be valid.  Analyze the
             aliquots according to the method beginning in Section  10.

       8.2.3  Calculate the average  percent recovery (R), and the standard deviation of the
             percent recovery (s),  for the results.  Wastewater background corrections must be
             made before R and s calculations are performed.

       8.2.4  Using the data from  Table  2, estimate the recovery and single-operator precision
             expected for the method, and compare these results to the values calculated in
             Section 8.2.3. If the data  are not comparable, review potential problem areas
             and repeat the test.

8.3     The analyst must calculate method performance criteria and define the performance of
       the laboratory  for each spike concentration and parameter being measured.

       8.3.1  Calculate upper and lower control limits for method performance as follows:

                    Upper Control Limit (UCL) = R +  3s
                    Lower Control Limit (1CL) = R - 3s

             where R and s are calculated as in Section 8.2.3.  The UCL and LCL  can be used
             to construct control  charts9 that are useful in observing trends in performance.

       8.3.2  The laboratory  must develop  and maintain  separate accuracy  statements of
             laboratory performance for wastewater samples. An accuracy statement for the
             method is defined as R  ± s. The accuracy statement should be developed by the
             analysis of four aliquots of wastewater as  described in Section 8.2.2,  followed by
             the calculation of R  and s. Alternatively, the analyst may use four wastewater
             data points gathered through the requirement for continuing quality control in
             Section 8.4. The accuracy statements should be updated regularly.9

8.4     The laboratory is required to collect in duplicate a portion of  their samples to monitor
       spike recoveries.  The frequency of spiked sample analysis must be at least 10% of all
       samples or one spiked  sample per month, whichever  is greater.  One aliquot of the
       sample must be spiked and analyzed as described in Section  8.2.  If the recovery for a
       particular parameter does not fall  within the control limits for  method performance, the
       results reported for that parameter in all samples processed as  part of the same set must
       be qualified as described in Section 13.3. The laboratory should monitor the frequency
       of data so qualified to ensure that it remains at or below 5%.

8.5     Before processing any samples, the analyst must  demonstrate  through the analysis of a
       1-L  aliquot of reagent water that all  glassware and reagent interferences are under
       control.  Each time a set of samples is extracted or there is a change in reagents, a
       laboratory  reagent blank  must  be processed  as  a  safeguard  against laboratory
       contamination.

-------
                                                                               Method 619
8.6     It is recommended that the laboratory adopt additional quality assurance practices for
       use with this method.  The specific practices that are most productive depend upon the
       needs of the laboratory and the nature of the samples.  Field duplicates may be analyzed
       to  monitor the precision of the  sampling technique.   When doubt  exists over the
       identification  of a peak on the chromatogram, confirmatory techniques such as gas
       chromatography  with  a  dissimilar  column,  specific element  detector,  or  mass
       spectrometer must be used.  Whenever possible, the laboratory should perform analysis
       of quality control materials and participate in relevant performance evaluation studies.

9.     SAMPLE COLLECTION, PRESERVATION, AND HANDLING

9.1     Grab samples must be collected in glass containers.  Conventional  sampling practices10
       should be followed; however, the bottle must not be  prerinsed  with sample before
       collection.   Composite samples should be collected in refrigerated glass containers in
       accordance with the requirements of the program. Automatic sampling equipment must
       be  as free as possible of plastic and other  potential sources of contamination.

9.2     The samples  must be  iced or refrigerated at 4°C from the  time of  collection until
       extraction.

9.3     All samples must be extracted within 7 days and completely analyzed within 40 days of
       extraction.

10.    SAMPLE EXTRACTION

10.1    Mark the water  meniscus on the  side of the sample bottle for later determination of
       sample volume.  Pour the entire sample into a 2-L separatory funnel.

10.2    Add  60 mL methylene chloride to  the sample bottle, seal, and shake 30 seconds to rinse
       the inner walls. Transfer the solvent to the separatory funnel and extract the sample by
       shaking the funnel for  two minutes with periodic  venting to release excess pressure.
       Allow the organic layer to separate from the water phase for a minimum of 10 minutes.
       If the emulsion interface between layers is more than one-third the volume of the solvent
       layer, the analyst must employ mechanical techniques to complete the phase separation.
       The optimum technique depends upon the sample, but may include stirring, filtration of
       the emulsion through glass wool, centrifugation, or other physical methods.  Collect the
       methylene chloride extract in a 250-mL Erlenmeyer flask.

10.3    Add  a second 60-mL volume of methylene chloride to the sample bottle and repeat the
       extraction procedure a second time,  combining the extracts in the Erlenmeyer flask.
       Perform a third extraction in the same manner.

10.4    Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10-mL concentrator tube
       to a 500-mL evaporative flask.  Other concentration devices or techniques may be used
       in place of the K-D if the requirements of Section 8.2 are met.

10.5    Pour the combined  extract through  a drying column  containing about 10 cm of
       anhydrous sodium sulfate, and collect the extract in the K-D  concentrator. Rinse the

-------
Method 619
       Erlenmeyer flask and column with 20- to 30-mL of methylene chloride to complete the
       quantitative transfer.

10.6    Add one or two clean  boiling chips to the evaporative flask and attach a three-ball
       Snyder column.  Prewet the Snyder column by adding about 1 mL methylene chloride
       to  the top.  Place the K-D apparatus on  a hot water bath, 60 to  65°C, so that the
       concentrator tube is partially immersed in the hot water, and the entire lower rounded
       surface  of the flask is  bathed with hot vapor.   Adjust the vertical position  of the
       apparatus and the water temperature as required to complete the concentration in 15 to
       20  minutes. At the proper rate of distillation, the balls of the column will actively  chatter
       but the chambers will not  flood with condensed solvent. When the apparent volume of
       liquid reaches 1 mL, remove the K-D apparatus and allow it to drain and cool for at least
       10  minutes.

10.7    Increase the temperature of the hot water bath to about 80°C.  Momentarily remove the
       Snyder column,  add 50  mL of hexane and a new boiling chip, and reattach the Snyder
       column.  Pour about 1 mL of hexane into the top  of the Snyder column and concentrate
       the solvent extract as before. Elapsed time of concentration should be 5 to 10 minutes.
       When the apparent volume of liquid reaches 1 mL, remove the K-D apparatus and allow
       it to drain and cool for at  least 10 minutes.

10.8    Remove the Snyder column and rinse the flask and its lower joint into the concentrator
       tube with 1 to 2 mL of hexane and adjust the volume to 10 mL.  A 5-mL  syringe is
       recommended for this operation.

       NOTE:  Precipitation  of triazines in the hexane may occur if the concentration in the original
       sample exceeded 500 ]ig/L.  If this occurs, redissolve the triazines in methylene chloride and
       analyze the extract using flame ionization gas chromatography. Stopper the concentrator tube and
       store refrigerated if further processing will not be performed immediately. If the extracts will be
       stored longer than two days, they should be transferred  to TFE-fluorocarbon-sealed screw-cap
       vials.  If the sample extract requires no  further cleanup, proceed with gas chromatographic
       analysis.  If the sample requires cleanup, proceed to Section 11.

10.9    Determine the original sample volume by refilling the sample bottle to the mark and
       transferring the water to a 1000-mL graduated cylinder.  Record the  sample volume to
       the nearest 5 mL.

11.    CLEANUP AND SEPARATION

11.1    Cleanup procedures may  not  be necessary for a relatively clean sample matrix.  The
       cleanup procedure  recommended in this method has  been used for the  analysis of
       various  industrial and municipal effluents.  If particular circumstances demand the use
       of  an alternative cleanup procedure, the analyst must determine the elution profile and
       demonstrate that the recovery of each compound of interest for the cleanup procedure
       is no less than 85%.

11.2    The following Florisil column cleanup procedure has been demonstrated to be applicable
       to  the nine  triazine pesticides listed in Table 3.

-------
                                                                               Method 619
       11.2.1  Add  a weight of  Florisil  (nominally  20  g)  predetermined  by  calibration
             (Section 7.4 and 7.5)  to a chromatographic column. Settle the Florisil by tapping
             the column.  Add anhydrous sodium sulfate to the top of the  Florisil to form a
             layer 1 to 2 cm deep. Add 60 mL of hexane to wet and rinse the sodium sulfate
             and Florisil.  Just prior to exposure of the sodium sulfate to  air, stop the elution
             of the hexane by closing the  stopcock on the chromatography  column. Discard
             the eluate.

       11.2.2  Adjust the sample extract volume to  10 mL with hexane and transfer it from the
             K-D concentrator tube to the Florisil column. Rinse the tube twice with 1 to 2 mL
             hexane, adding each rinse to the column.

       11.2.3  Drain the column until the sodium  sulfate layer is nearly  exposed.  Elute the
             column with 200 mL of 6% (v/v) ethyl ether in hexane (Fraction 1) using a drip
             rate of about 5 mL/min.  This fraction may be discarded.  Place a 500-mL K-D
             flask and clean concentrator tube under the chromatography column. Elute the
             column into the flask, using 200 mL of 15% (v/v) ethyl ether in hexane (Fraction
             2).  Perform  a third elution  using 200 mL of 50% (v/v) ethyl ether in hexane
             (Fraction 3), and a final elution with 200 mL of 100% ethyl ether (Fraction 4), into
             separate K-D flasks.  The elution patterns for nine of the pesticides are shown in
             Table 3.

       11.2.4  Concentrate the eluates by standard K-D techniques (Section 10.6), substituting
             hexane for the glassware rinses and using the water bath at  about 85°C. Adjust
             final volume to  10 mL with hexane.  Analyze by gas chromatography.

12.    GAS CHROMATOGRAPHY

12.1    Table 1 summarizes  the recommended operating conditions for the  gas chromatograph.
       Included in this table are estimated retention times and method detection limits that can
       be achieved by this  method.  An example of the separation achieved by Column  1 is
       shown in Figure 1. Other packed columns, chromatographic conditions, or detectors may
       be used if the requirements of Section 8.2 are met.  Capillary (open-tubular) columns may
       also be used if the relative standard deviations of responses for replicate injections are
       demonstrated to be less than 6% and the requirements of Section 8.2 are met.

12.2    Calibrate the system daily as described in Section 7.

12.3    If the internal standard approach is being used, add the internal  standard to sample
       extracts immediately before injection into the instrument.  Mix thoroughly.

12.4    Inject 1 to 5 uL of the sample extract using the solvent-flush technique.11   Record the
       volume injected to the nearest 0.05 uL, and the resulting  peak size in area or peak height
       units.  An automated system that consistently injects a constant volume of extract may
       also be used.

12.5    The width of the retention-time window used to make  identifications should be based
       upon measurements of actual retention-time variations  of standards over the course of
       a  day.  Three times the standard deviation of a retention time can  be used to calculate

-------
Method 619
       a suggested window  size for a compound.  However, the experience of the analyst
       should weigh heavily in the interpretation of chromatograms.

12.6    If the response for the peak exceeds the working range of the system, dilute the extract
       and reanalyze.

12.7    If the measurement of the peak response is prevented by the presence of interferences,
       further cleanup is required.

13.    CALCULATIONS

13.1    Determine the concentration of individual compounds in the sample.

       13.1.1  If the external standard calibration procedure is used,  calculate the amount of
              material injected from the peak response using the calibration curve or calibration
              factor in Section 7.2.2.  The concentration in the sample  can  be calculated as
              follows:


                                       Equation 2


                                                    (4)  (V)
                             Concentration, \ig/L -
              where
              A  = Amount of material injected, in ng
              Vt  = Volume of extract injected, in uL
              Vt  = Volume of total extract, in uL
              V.  = Volume of water extracted, in mL
                                                      ) (V)
       13.1.2  If  the  internal  standard  calibration  procedure  was used,  calculate  the
              concentration in the sample using the response factor (RF) determined in Section
              7.3.2 as follows:

-------
                                                                               Method 619
                                       Equation 3


                         Concentration, \ig/L -
                                                (At)  (RF)
             where
             As  = Response for parameter to be measured
             Ais  = Response for the internal standard
             Is   = Amount of internal standard added to each extract, in ug
             V0  = Volume of water extracted, in L
13.2    Report results in  micrograms per  liter without correction for recovery data.  When
       duplicate and spiked samples are analyzed, report all data obtained with the sample
       results.

13.3    For samples processed as part of a set where the laboratory spiked sample recovery falls
       outside of the control limits in Section 8.3, data for the affected parameters must be
       labeled as suspect.

14.    GC/MS CONFIRMATION

14.1    It is  recommended  that  GC/MS  techniques be judiciously employed  to  support
       qualitative compound identifications made with this method. The mass spectrometer
       should be capable of scanning the mass range from 35 amu to a mass 50 amu above the
       molecular weight  of the compound. The instrument must be capable of scanning the
       mass range at a rate to produce at least 5 scans per peak but not to exceed 7 seconds per
       scan utilizing a 70 V (nominal) electron energy in the electron impact ionization mode.
       A GC-to-MS interface constructed of all glass or glass-lined materials is recommended.
       A computer system should be  interfaced to the  mass spectrometer  that allows the
       continuous  acquisition and storage on machine-readable  media  of all  mass spectra
       obtained throughout the duration of the chromatographic program.

14.2    Gas chromatographic columns and conditions should be selected for optimum separation
       and performance.   The conditions selected must be compatible with standard GC/MS
       operating practices. Chromatographic tailing factors of less than 5.0 must be achieved.12

14.3    At the beginning of each day that confirmatory analyses are to be performed,  the GC/MS
       system must be  checked  to see  that  all  decafluorotriphenyl  phosphine  (DFTPP)
       performance criteria are achieved.13

14.4    To confirm an identification of a compound, the background-corrected mass spectrum
       of the compound must be obtained from the sample extract and compared with a  mass
       spectrum from a stock or calibration standard analyzed under the same chromatographic

-------
Method 619
       conditions.  It is recommended that at least 25 ng of material be injected into the GC/MS.
       The criteria below must be met for qualitative confirmation.

       14.4.1 All ions that are present above 10% relative abundance in the mass spectrum of
             the standard must be present in the mass spectrum of the sample with agreement
             to ±10%.  For example, if the relative abundance of an ion is 30% in the mass
             spectrum of the standard, the allowable limits for the relative abundance of that
             ion in the mass spectrum for the sample would be 20 to 40%.

       14.4.2 The retention time of the compound in the sample must be within 6 seconds of
             the same compound in the standard solution.

       14.4.3 Compounds that have very similar mass spectra can be explicitly identified by
             GC/MS only on the basis of retention time data.

14.5    Where available, chemical ionization mass spectra may be employed to aid in the
       qualitative identification process.

14.6    Should these MS procedures fail to provide satisfactory results, additional steps may be
       taken before reanalysis. These may include the use of alternate packed or capillary GC
       columns or additional cleanup (Section 11).

15.    METHOD PERFORMANCE

15.1    The method  detection limit (MDL) is defined as the minimum  concentration  of  a
       substance that can be measured and reported with 99% confidence that the value is above
       zero.14 The MDL concentrations listed in Table 1 were estimated from the response of
       the thermionic bead nitrogen detector to each compound.  The estimate is based upon
       the amount of material required to yield a signal  5 times the GC background noise,
       assuming a 5-uL injection from a 10-mL final extract of a 1-L sample.

15.2    In a single  laboratory (either West Cost  Technical Services, Inc., or Midwest Research
       Institute), using effluents from pesticide  manufacturers and publicly owned treatment
       works (POTW), the average recoveries presented in Table  2 were obtained after Florisil
       cleanup.1'2 The standard deviations of the percent recoveries of these measurements are
       also included in Table 2.

-------
                                                                               Method 619
                                    References

1.      "Pesticide Methods Evaluation," Letter Report #11 for EPA Contract No. 68-03-2697.
       Available from U.S. Environmental Protection Agency, Environmental Monitoring and
       Support Laboratory, Cincinnati, Ohio  45268.

2.      "Development of Analytical  Test Procedures for Organic Pollutants in Wastewater—
       Application to Pesticides," EPA Report 600/4-81-017, U.S. Environmental Protection
       Agency, Cincinnati, Ohio.   PB#82  132507, National  Technical  Information  Service,
       Springfield, Virginia.

3.      "Methods for Benzidine, Chlorinated  Organic Compounds,  Pentachlorophenol and
       Pesticides  in  Water  and   Wastewater,"  U.S.  Environmental  Protection  Agency,
       Environmental Monitoring and Support Laboratory - Cincinnati, Ohio, September 1978.

4.      ASTM Annual Book of Standards, Part 31, D3694, "Standard Practice for Preparation of
       Sample Containers and for Preservation," American Society for Testing and Materials,
       Philadelphia, Pennsylvania, p. 679, 1980.

5.      "Carcinogens—Working  with  Carcinogens," Department of Health, Education, and
       Welfare, Public Health  Service,  Center for Disease Control, National Institute for
       Occupational Safety and Health, Publication No. 77-206, August 1977.

6.      "OSHA Safety and Health Standards,  General Industry" (29 CFR 1910), Occupational
       Safety and Health Administration, OSHA 2206 (Revised, January  1976).

7.      "Safety in Academic Chemistry Laboratories,"  American Chemical Society Publication,
       Committee on Chemical  Safety, 3rd Edition, 1979.

8.      ASTM Annual Book of  Standards,  Part 31, D3086, Appendix X3, "Standardization of
       Florisil Column by Weight Adjustment Based on Adsorption of Laurie Acid," American
       Society for Testing and Materials, Philadelphia, Pennsylvania, p. 765, 1980.

9.      "Handbook for  Analytical Quality Control in Water  and Wastewater Laboratories,"
       EPA-600/4-79-019, U. S.  Environmental Protection Agency, Environmental Monitoring
       and Support Laboratory  - Cincinnati, Ohio, March 1979.

10.     ASTM  Annual Book of Standards,  Part 31, D3370, "Standard Practice for Sampling
       Water,"  American Society for Testing and Materials, Philadelphia, Pennsylvania, p. 76,
       1980.

11.     Burke,  J.A.    "Gas Chromatography for Pesticide Residue Analysis;  Some Practical
       Aspects," Journal of the Association of Official Analytical Chemists, 48, 1037  (1965).

12.     McNair, H.M., and Bonelli, E.J. Basic Chromatography, Consolidated Printing, Berkeley,
       California, p. 52, 1969.

-------
Method 619
13.     Eichelberger, J.W., Harris, L.E., and Budde, W.L.  "Reference Compound to Calibrate Ion
       Abundance  Measurement  in Gas Chromatography-Mass Spectrometry," Analytical
       Chemistry, 47, 995 (1975).

14.     Glaser, J.A. et al.  "Trace Analysis for Wastewaters," Environmental Science & Technology,
       15, 1426 (1981).

-------
                                                                             Method 619
Table 1.      Chromatographic Conditions and Method Detection Limits

                                 Retention Time (min)          Method
                             	1	  Detection Limit
          Parameter             Column 1     Column 2         (ug/L)
          Prometon                6.9           4.9                0.03
          Atraton                  -            6.3               ND
          Propazine                9.2           6.7                0.03
          Terbuthylazine           10.2           7.3                0.03
          Secbumeton              _            8.3              ND
          Atrazine                 12.4           9.4                0.05
          Prometryn               13.8          10.3                0.06
          Terbutryn               15.4           _                 0.05
          Simazine                16.3          12.7                0.06
          Ametryn                17.7          14.0                0.06
          Simetryn                23.0           _                 0.07

ND = Not determined

Column 1 conditions: Supelcoport (80 / 100 mesh) coated with 5% Carbowax 20M-TPA packed
in a glass column 1.8 m long by 2 mm ID with helium carrier gas at a flow rate of 30 mL/min.
Column  temperature,  isothermal at 200°C. A thermionic bead detector was used with this
column to determine the MDL.

Column 2 conditions: Gas Chrom Q (100 / 120 mesh) coated with 1.0% Carbowax 20 M packed
in a glass column 1.8  m long by 4 mm ID with helium carrier gas at 80 mL/min flow rate.
Column temperature, isothermal at  155°C.

-------
Method 619
Table 2. Single-Laboratory Accuracy and

Parameter
Ametryn
Atrazine

Prometon

Prometryn
Propazine
Simatryn
Simazine

Terbuthylazine
Terbutryn

Sample
Type*
3
3
3
1
1
2
3
3
1
3
3
3
1
3
3
3
1
2

Spike
(fig/L)
4,000
2,000
300
1,000
130
260
2,000
50
516
15
30
15
115
10
100
15
968
169
Precision
Number
of
Replicates
2
2
2
7
7
7
2
2
7
2
2
7
2
2
2
7
7

Mean
Recovery
(%)
104
118
108
177
67
51
76
110
54
116
183
182
152
99
114
100
83
89

Standard
Deviation
/o/ \
\'0>
—

15.2
3.9
3.0
-
6.5
-
24.3

—
10.0
24.0
*Sample Type:
    1 = Industrial process water
    2 = Industrial effluent
    3 = 80% Industrial process water / 20% industrial effluent

-------
                                                                                Method 619
Table 3.       Florisil Fractionation Patterns

         Parameter
         Propazine
         Terbuthylazine
         Atrazine
         Ametryn
         Prometryn
         Simazine
         Atraton
         Secbumeton                                                      100
         Prometon                                                        100

Florisil eluate composition by fraction:
       Fraction  1 = 200 mL of 6% ethyl ether in hexane
       Fraction  2 = 200 mL of 15%  ethyl ether in hexane
       Fraction  3 = 200 mL of 50%  ethyl ether in hexane
       Fraction  4 = 200 mL of ethyl ether
Percent Recovery by Fraction
No. 1
0
0
0




No. 2
90
30
20




No. 3
10
70
80
100
100
100
100
No. 4








-------
Method 619
                               i Prometon
                                    /Propazine
                                  I    /Terbuthylazine
^Atrazine

 Prometryn

       /
                                            /     /Simazine
                     i     i     I     1     I    i     i     i     i     i
                         5.0       10.0      15.0      20.0      25.0
                              Retention Time (minutes)
                                                                             A52-002-49A
        Figure 1.    Gas Chromatogram of Triazine Pesticides on Column 1
                    For Conditions, See Table 1

-------