SEPA
United States
Environmental Protection
Agency
 www.epa.gov
  Promulgated 1984
   Method 613: 2,3,7,8-Tetrachloro-
   dibenzo-p-Dioxin

-------
                              APPENDIX A TO PART 136
       METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND
                            INDUSTRIAL WASTEWATER

              METHOD 613—2,3,7,8-TETRACHLORODIBENZO-P-DIOXIN
1.      Scope and Application

1.1     This method covers the determination of 2,3,7,8-tetrachlorodibenzo-p-dioxin
       (2,3,7,8-TCDD).  The following parameter may be determined by this method:

237 8-TCDD

Parameter


STORET No.
34675

CAS No.
1746-01-6

1.2     This is a gas chromatographic/mass spectrometer (GC/MS) method applicable to the
       determination of 2,3,7,8-TCDD in municipal and industrial discharges as provided
       under 40 CFR Part 136.1.  Method 625 may be used to screen samples for
       2,3,7,8-TCDD. When the screening test is positive, the final qualitative confirmation
       and quantification must be made using Method 613.

1.3     The method detection limit (MDL, defined in Section 14.1)1 for 2,3,7,8-TCDD is listed
       in Table 1.  The  MDL for a specific wastewater may be different from that listed,
       depending upon the nature of interferences in the sample matrix.

1.4     Because of the extreme toxicity of this compound, the analyst  must prevent exposure
       to himself, of to others, by materials knows or believed to contain 2,3,7,8-TCDD.
       Section 4 of this method contains guidelines and protocols that serve as minimum
       safe-handling standards in a limited-access laboratory.

1.5     Any modification of this method, beyond those  expressly permitted, shall be
       considered as a  major modification subject to application and  approval of alternate
       test procedures under 40 CFR Parts 136.4 and 136.5.

1.6     This method is restricted to use by or under the supervision of analysts experienced
       in the use of a gas chromatograph/mass spectrometer and in the interpretation of
       mass spectra. Each analyst must demonstrate the ability to generate acceptable results
       with this method using the procedure described in Section 8.2.

2.      Summary of Method

2.1     A measured volume of sample, approximately 1 L, is spiked with an internal standard
       of labeled 2,3,7,8-TCDD and extracted with methylene chloride using a separatory
       funnel.  The methylene chloride  extract is exchanged to hexane during concentration
       to a volume of 1.0 mL or less. The extract is then analyzed by capillary column
       GC/MS to separate and measure 2,3,7,8-TCDD.2,3

-------
2.2     The method provides selected column chromatographic cleanup proceudres to aid in
       the elimination of interferences that may be encountered.

3.      Interferences

3.1     Method interferences may be caused by contaminants in solvents, reagents, glassware,
       and other sample processing hardware that lead to discrete artifacts and/or elevated
       backgrounds at the masses (m/z) monitored.  All of these materials must be routinely
       demonstrated to  be free from interferences under the conditions of the analysis by
       running laboratory reagent blanks as described in Section 8.1.3.

       3.1.1    Glassware must be scrupulously cleaned.4  Clean all glassware as soon as
              possible after  use by rinsing with the last solvent used in  it.  Solvent rinsing
              should  be followed by detergent washing with hot water, and rinses with tap
              water and distilled water.  The glassware should then be drained dry, and
              heated  in a muffle furnace at 400°C for 15-30 minutes.  Some thermally stable
              materials, such as PCBs, may not be eliminated by the treatment.  Solvent
              rinses with acetone and pesticide quality hexane may be substituted for the
              muffle furnace heating. Thorough rinsing with such solvents usually
              eliminates PCB interference.  Volumetric ware should not be heated in a muffle
              furnace. After drying and cooling, glassware should be sealed and stored in a
              clean environment to prevent any accumulation of dust or other contaminants.
              Store inverted or capped with aluminum foil.

       3.1.2    The use of high purity reagents and solvents helps to mininmize interference
              problems. Purification of solvents by distillation in all-glass systems may be
              required.

3.2     Matrix interferences may be caused by contaminants that are coextracted  from the
       sample. The extent of matrix interferences will vary considerably from source to
       source, depending upon the nature and diversity of the industrial complex or
       municipality being sampled.  2,3,7,8-TCDD  is often associated with other  interfering
       chlorinated compounds which are at concentrations several magnitudes higher than
       that of 2,3,7,8-TCDD.  The cleanup producers in Section  11 can be used to overcome
       many of these  interferences, but unique samples may require additional cleanup
       approaches1'5 7  to eliminate false positives and achieve the MDL listed in Table 1.

3.3     The primary column, SP-2330 or equivalent, resolves  2,3,7,8-TCDD from the other 21
       TCDD insomers.  Positive results  using any other gas chromatographic  column must
       be confirmed using the primary column.

4.      Safety

4.1     The toxicity or carcinogenicity of each reagent used in this method has not been
       precisely defined; however, each chemical compound should be treated as a potential
       health hazard.  From this viewpoint, exposure to these chemicals must be reduced to
       the lowest possible level by whatever means available. The laboratory is  responsible
       for maintaining a current awareness file of OSHA regulations regarding the safe
       handling of the chemicals specified in this method. A reference file of material data
       handling sheets should also be made available to all personnel involved in the

-------
       chemical analysis.  Additional references to laboratory safety are available and have
       been identified810 for the information of the analyst. Benzene and 2,3,7,8-TCDD have
       been identified as suspected human or mammalian carcinogens.

4.2     Each laboratory must develop a strict safety program for handling 2,3,7,8-TCDD. The
       following laboratory practices are recommended:

       4.2.1  Contamination of the laboratory will be minimized by conducting all
             manipulations in a hood.

       4.2.2  The effluents of sample splitters for the gas chromatograph and roughing
             pumps on the GC/MS should pass through either a column of activated
             charcoal or  be bubbled through a trap containing oil or high-boiling alcohols.

       4.2.3  Liquid waste should be dissolved in methanol or ethanol and irradiated with
             ultraviolet light with a wavelength greater than 290 nm for several days.  (Use
             F 40 BL lamps or equivalent). Analyze liquid wastes and dispose of the
             solutions when  2,3,7,8-TCDD can no longer be detected.

4.3     Dow Chemical U.S.A. has issued the following precautimns (revised November 1978)
       for safe handling of 2,3,7,8-TCDD in the laboratory:

       4.3.1  The following statements  on safe handling are as complete as possible on the
             basis of available toxicological information. The precautions  for safe handling
             and use are necessarily general in nature since detailed, specific
             recommendations can be made only for the particular exposure and
             circumstances of each individual use.  Inquiries about specific operations or
             uses may be addressed to the Dow Chemical Company. Assistance in
             evaluating the health hazards of particular plant conditions may be obtained
             from certain consulting laboratories and from State Departments of Health or
             of Labor, many  of which have an industrial health service. 2,3,7,8-TCDD is
             extremely toxic  to laboratory animals.  However, it has been handled for years
             without injury in analytical and biological  laboratories.  Techniques used in
             handling radioactive and  infectious materials are applicable to 2,3,7,8,-TCDD.

             4.3.1.1 Protective equipment—Throw-away plastic gloves, apron or lab coat,
                    safety glasses, and a lab hood adequate  for radioactive work.

             4.3.1.2 Training—Workers must be trained in the proper method of removing
                    contaminated gloves and clothing without contacting the exterior
                    surfaces.

             4.3.1.3 Personal hygiene—Thorough washing of hands and forearms after each
                    manipulation and  before breaks (coffee,  lunch, and shift).

             4.3.1.4 Confinement—Isolated work area, posted with signs, segregated
                    glassware and tools, plastic-backed  absorbent paper on benchtops.

-------
 4.3.1.5 Waste—Good technique includes minimizing contaminated waste.
       Plastic bag liners should be used in waste cans. Janitors must be
       trained in the safe handling of waste.

 4.3.1.6 Disposal of wastes—2,3,7,8-TCDD decomposes above 800°C.  Low-level
       waste such as absorbent paper, tissues, animal remains, and plastic
       gloves may be burned in a good incinerator.  Gross quantities
       (milligrams) should be packaged securely and disposed through
       commercial or governmental  channels which are capable of handling
       high-level radioactive wastes or extremely toxic wastes.  Liquids should
       be allowed to evaporate in a  good hood and in  a disposable container.
       Residues may then be handled  as above.

 4.3.1.7 Decontamination—For personal decontamination, use any mild  soap
       with plenty of scrubbing action.  For decontamination of glassware,
       tools, and surfaces, Chlorothene NU Solvent (Trademark of the  Dow
       Chemical Company) is the least toxic solvent shown to be effective.
       Satisfactory cleaning may be  accomplished by rinsing with Chlorothene,
       then washing with any detergent and water.  Dishwater may  be
       disposed to the sewer. It  is prudent to minimize solvent wastes
       because they may require special disposal through commercial sources
       which are expensive.

 4.3.1.8 Laundry—Clothing known to be  contaminated should be disposed with
       the precautions described under Section 4.3.1.6. Lab coats or  other
       clothing worn in 2,3,7,8-TCDD work areas may be laundered.

       Clothing should be collected  in plastic bags. Persons who convey the
       bags and launder the clothing should be advised of the hazard and
       trained in proper handling. The clothing may be put into a washer
       without contact if the launderer knows the problem. The washer
       should be run through a cycle before being used again for other
       clothing.

 4.3.1.9 Wipe tests—A useful method of determining cleanliness of work
       surfaces and  tools is to wipe  the surface with a  piece of filter  paper.
       Extraction and analysis by gas chromatography can achieve a limit of
       sensitivity of 0.1 ug per wipe. Less than 1 ug of 2,3,7,8-TCDD per
       sample indicates  acceptable cleanliness; anything higher warrants
       further cleaning.  More than  10 ug on a wipe sample constitutes an
       acute hazard and requires prompt cleaning before  further use of the
       equipment or work space.  A high (10 ug) 2,3,7,8-TCDD level indicates
       that unacceptable work practices  have been employed in the past.

4.3.1.10 Inhalation—Any  procedure that may produce airborne contamination
       must be done with good ventilation. Gross losses  to a ventilation
       system must not  be allowed.  Handling of the dilute solutions normally
       used in analytical and animal work presents no inhalation hazards
       except in the case of an accident.

-------
             4.3.1.11 Accidents—Remove contaminated clothing immediately, taking
                    precautions not to contaminate skin or other articles.  Wash exposed
                    skin vigorously and repeatedly until medical attention is obtained.

5.      Apparatus and Materials

5.1     Sampling equipment, for discrete or composite sampling.

       5.1.1  Grab sample bottle—1 L or 1 qt, amber glass, fitted with a screw cap lined with
             Teflon. Foil may be substituted for Teflon if the sample is not corrosive.  If
             amber bottles are not available, protect samples from light.  The bottle and cap
             liner must be washed, rinsed with acetone or methylene chloride, and dried
             before use to minimize contamination.

       5.1.2  Automatic sampler (optional)—The sampler must incorporate glass sample
             containers for the  collection of a minimum of 250  mL of sample. Sample
             containers must be kept refrigerated at 4°C and protected from light during
             compositing.  If the sampler uses a peristaltic pump, a minimum length of
             compressible silicone rubber tubing may be used.  Before use, however, the
             compressible tubing should be thoroughly rinsed  with methanol, followed by
             repeated  rinsings with distilled water to minimize the potential for
             contamination of the  sample. An integrating  flow meter is required to collect
             flow proportional composites.

       5.1.3  Clearly label all samples as "POISON"  and ship according to U.S. Department
             of Transportation regulations.

5.2     Glassware (All specifications are suggested.  Catalog numbers are included for
       illustration only)

       5.2.1  Separatory funnels—2 L and 125 mL, with Teflon  stopcock.

       5.2.2  Concentrator tube, Kuderna-Danish—10 mL, graduated (Kontes K-570050-1025
             or equivalent). Calibration must be checked at the volumes employed in the
             test.  Ground glass stopper is used to prevent evaporation of extracts.

       5.2.3  Evaporative flask, Kuderna-Danish—500 mL (Kontes K-570001-0500 or
             equivalent).  Attach to concentrator tube with springs.

       5.2.4  Snyder column, Kuderna-Danish—Three-ball  macro (Kontes K-503000-0121 or
             equivalent).

       5.2.5  Snyder column, Kuderna-Danish—Two-ball micro (Kontes K-569001-0219 or
             equivalent).

       5.2.6  Vials—10-15 mL, amber glass, with Teflon-lined screw cap.

       5.2.7  Chromatographic  column—300 mm long x 10 mm ID, with Teflon stopcock and
             coarse frit filter disc at bottom.

-------
       5.2.8  Chromatographic column—400 mm long x 11 mm ID, with Teflon stopcock and
             coarse frit filter disc at bottom.

5.3     Boiling chips—Approximately 10/40 mesh. Heat to 400°C for 30 minutes or Soxhlet
       extract with methylene chloride.

5.4     Water bath—Heated, with concentric ring cover, capable of temperature control
       (±2°C).  The bath should be used in a hood.

5.5     GC/MS system

       5.5.1  Gas chromatograph—An analytical system complete with a temperature
             programmable gas chromatograph and all required accessories including
             syringes, analytical columns, and gases. The injection port must be designed
             for capillary columns. Either split, splitless, or on-column injection techniques
             may be employed, as long as the requirements of Section 7.1.1  are achieved.

       5.5.2  Column—60 m long x 0.25 mm ID glass or fused silica, coated with SP-2330 (or
             equivalent) with a film thickness of 0.2 um. Any equivalent column must
             resolve 2, 3, 7, 8-TCDD from the other 21 TCDD isomers.16

       5.5.3  Mass spectrometer—Either a low resolution mass spectrometer (LRMS) or a
             high resolution mass spectrometer (HRMS) may be used. The  mass
             spectrometer must be equipped with a 70 V (nominal) ion source and be
             capable of aquiring m/z abundance data in real time selected ion monitoring
             (SIM)  for groups of four or more masses.

       5.5.4  GC/MS interface—Any GC to MS interface can be used that achieves the
             requirements of Section 7.1.1. GC to MS interfaces constructed of all glass or
             glass-lined materials are recommended. Glass surfaces can  be  deactivated by
             silanizing with dichlorodimethylsilane.  To achieve maximum sensitivity,  the
             exit end of the capillary column should be placed in the  ion source.  A short
             piece of fused silica capillary can be used as the interface to overcome problems
             associated with straightening the exit end of glass capillary  columns.

       5.5.5  The SIM data acquired during the chromatographic program is defined as the
             Selected Ion Current Profile (SICP).  The SICP can be acquired under computer
             control or as a real time analog output.  If computer control is  used, there must
             be software available to plot the SICP and report peak height or  area data for
             any m/z in the SICP between specified time or scan number limits.

5.6     Balance—Analytical, capable of accurately weighing 0.0001 g.

6.      Reagents

6.1     Reagent water—Reagent water is defined as a water in which an interferent is not
       observed at the MDL of 2, 3, 7, 8-TCDD.

-------
6.2     Sodium hydroxide solution (10 N)—Dissolve 40 g of NaOH (ACS) in reagent water
       and dilute to 100 mL.  Wash the solution with methylene chloride and hexane before
       use.

6.3     Sodium thiosulfate—(ACS) Granular.

6.4     Sulfuric acid—Concentrated (ACS, sp. gr. 1.84).

6.5     Acetone, methylene chloride, hexane, benzene, ortho-xylene, tetradecane—Pesticide
       quality or equivalent.

6.6     Sodium sulfate—(ACS) Granular, anhydrous.  Purify  by heating at 400°C for four
       hours in a shallow tray.

6.7     Alumina—Neutral, 80/200 mesh  (Fisher Scientific Co., No. A-540 or equivalent).
       Before use, activate for 24 hours at 130°C in a foil-covered glass container.

6.8     Silica gel—High purity grade, 100/120 mesh (Fisher Scientific Co., No. S-679 or
       equivalent).

6.9     Stock standard solutions (1.00 ug/uL)—Stock standard solutimns can be prepared
       from pure standard materials or purchased as certified solutions.  Acetone should be
       used as the solvent for spiking solutions; ortho-xylene is recommended for calibration
       standards for split injectors;  and tetradecane is recommended for splitless or on-colum
       injectors. Analyze stock internal standards to verify the absence of native
       2,3,7,8-TCDD.

       6.9.1  Prepare stock standard solutions  of 2,3,7,8-TCDD (mol wt 320) and either 37C14
             2,3,7,8-TCDD (mol wt 328)  or 13KC112K 2,3,7,8-TCDD (mol wt 332) in an
             isolated area by accurately weighing about 0.0100 g of pure material.  Dissolve
             the material in pesticide  quality solvent and dilute  to volume  in a 10 mL
             volumetric flask. When  compound purity is assayed to be 96% or greater, the
             weight can be  used without correction to  calculate the concentration of the
             stock standard.  Commercially  prepared stock standards can be used at any
             concentration if they are certified by the manufacturer or by an independent
             source.

       6.9.2  Transfer the stock standard solutions into Teflon-sealed screw-cap bottles. Store
             in an isolated refrigerator protected from  light.   Stock standard solutions should
             be checked frequently for signs of degradation  or evaporation, especially just
             prior to preparing calibration standards or spiking  solutions from them.

       6.9.3  Stock standard solutions must be replaced after six months, or sooner  if
             comparison with check standards indicates a problem.

6.10    Internal standard spiking solution (25 ng/mL)—Using stock standard solution,
       prepare a spiking solution in acetone of either 13KC112K or 37KC14K 2,3,7,8-TCDD at a
       concentration of 25 ng/mL.  (See Section 10.2)

6.11    Quality control check sample concentrate—See Section 8.2.1.

-------
7.      Calibration

7.1     Establish gas chromatograhic operating conditions equivalent to those given in Table 1
       and SIM conditions for the mass spectrometer as described in Section 12.2. The
       GC/MS system must be calibrated using the internal standard technique.

       7.1.1  Using stock standards, prepare calibration standards that will allow
             measurement of relative response factors of at least three concentration ratios of
             2,3,7,8-TCDD to internal standard. Each calibration standard must be prepared
             to contain the internal standard at a concentration of 25 ng/mL. If any
             interferences are contributed by the internal standard at m/z 320 and 322, its
             concentration may be reduced in the calibration standards and in the internal
             standard spiking solution (Section 6.10).  One of the calibration standards
             should contain 2,3,7,8-TCDD at a  concentration near, but above, the MDL and
             the other 2,3,7,8-TCDD concentrations should correspond to the expected range
             of concentrations found in real samples or should define the working range of
             the GC/MS system.

       7.1.2  Using injections of 2-5 uL, analyze each calibration standardaccording to
             Section 12 and tabulate peak height or  area response against the concentration
             of 2,3,7,8-TCDD and internal standard.  Calculate response  factors  (RF) for
             2,3,7,8-TCDD using Equation 1.

                                       Equation 1

                                    RF
                                          (Als) (C.)


             where:
                    As = SIM response for 2,3,7,8-TCDD m/z 320.
                    Ais = SIM response for the internal standard, m/z 332 for 13jf
                         2,3,7,8-TCDD m/z 328 for 37C14 2,3,7,8-TCDD.
                    Cis = Concentration of the internal standard (ug/L).
                    Cs = Concentration of 2,3,7,8-TCDD (ug/L).

             If the RF value over the working range is a constant (<10% relative standard
             deviation, RSD),  the RF can be assumed to be invariant and the average RF can
             be used for calculations.  Alternatively, the results can  be used to plot a
             calibration curve of response  ratios, As/Ais, vs. concentration  ratios Cs/Cis*.

       7.1.3  The working calibration curve or RF must be verified on each working day by
             the measurement of one or more 2,3,7,8-TCDD calibration standards. If the
             response for 2,3,7,8-TCDD varies from the predicted response by more  than
       *This equation corrects an error made in the original method publication (49 FR 43234,
October 26, 1984). This correction will be formalized through a rulemaking in FY97.

-------
             ±15%, the test must be repeated using a fresh calibration standard.
             Alternatively, a new calibration curve must be prepared.

7.2     Before using any cleanup procedure, the analyst must process a series of calibration
       standards through the procedure to validate elution patterns and the absence of
       interferences from  the reagents.

8.      Quality Control

8.1     Each laboratory that uses this method is required to operate a formal quality control
       program. The minimum requirements of this program consist of an initial
       demonstration of laboratory capability and an  ongoing analysis of spiked samples to
       evaluate and document data quality. The  laboratory must maintain records to
       document the quality of data that is generated. Ongoing data quality checks are
       compared with established performance criteria to determine if the results of analyses
       meet the performance characteristics of the method.  When results of sample spikes
       indicate atypical method performance, a quality control check standard must be
       analyzed to confirm that the measurements were performed in an in-control mode of
       operation.

       8.1.1  The analyst must make an initial, one-time, demonstration of the ability to
             generate acceptable  accuracy and precision with this  method. This ability is
             established as described in Section 8.2.

       8.1.2  In recognition of advances that are occurring in chromatography, the analyst is
             permitted certain options (detailed in Sections  10.5, 11.1, and 12.1) to improve
             the separations or lower the cost of  measurements. Each time such a
             modification is made to the method, the analyst is required to repeat the
             procedure in Section 8.2

       8.1.3  Before processing any samples, the analyst must analyze a reagent water  blank
             to demonstrate that interferences from the analytical  system and glassware are
             under control. Each time a set of samples is extracted or reagents are changed,
             a  reagent water blank must be processed as a safeguard against laboratory
             contamination.

       8.1.4  The laboratory must, on an ongoing basis, spike and analyze a minimum of
             10% of all samples with native 2,3,7,8-TCDD to monitor and evaluate laboratory
             data quality.  This procedure  is described in Section 8.3.

       8.1.5  The laboratory must, on an ongoing basis, demonstrate through the analyses of
             quality control check standards that the  operation of the measurement system is
             in control. This procedure is  described in Section 8.4. The frequency of the
             check standard analyses is equivalent to 10% of all samples analyzed but may
             be reduced if spike recoveries from  samples (Section  8.3) meet all specified
             quality control criteria.

       8.1.6  The laboratory must maintain performance records to document the quality of
             data that is generated.  This procedure is described in Section 8.5.

-------
8.2     To establish the ability to generate acceptable accuracy and precision, the analyst must
       perform the following operations.

       8.2.1  A quality control (QC) check sample concentrate is required containing
             2,3,7,8-TCDD at a concentration of 0.100 ug/mL in acetone.  The QC check
             sample concentrate must be obtained from the U.S. Environmental Protection
             Agency, Environmental Monitoring and Support Laboratory in Cincinnati,  Ohio,
             if available.  If not available from that source, the QC check sample concentrate
             must be obtained from another external source. If not available from either
             source above, the QC check sample concentrate must be prepared by the
             laboratory using stock standards prepared independently from those used  for
             calibration.

       8.2.2  Using a pipet, prepare QC check samples at a concentration of 0.100 l-ig/L
             (100 ng/L) by adding 1.00 mL of QC check sample concentrate to each of four
             1 L aliquots  of reagent water.

       8.2.3  Analyze the well-mixed QC check samples according to the method beginning
             in Section 10.

       8.2.4  Calculate the average recovery (X) in l-ig/L, and the standard deviation of the
             recovery (s)  in l-ig/L, for 2,3,7,8-TCDD using the four results.

       8.2.5  Compare s and (X)  with the corresponding acceptance criteria for precision and
             accuracy, respectively, found in Table 2.  If s and X meet the acceptance
             criteria, the system performance is acceptable and  analysis of actual samples
             can begin. If s exceeds the precision limit or  X falls outside the range for
             accuracy, the system performance is  unacceptable  for 2,3,7,8-TCDD.  Locate and
             correct the source of the problem and repeat the test beginning with
             Section 8.2.2.

8.3     The laboratory must, on an ongoing basis, spike at least  10% of the samples from each
       sample site being monitored to assess accuracy.  For laboratories analyzing one to ten
       samples per month, at  least one spiked sample per month is  required.

       8.3.1  The concentration of the spike in the sample should be determined as follows:

             8.3.1.1  If, as in compliance  monitoring, the concentration of 2,3,7,8-TCDD  in
                    the sample is being  checked against a regulatory concentration limit,
                    the spike should be at that limit or one to  five times higher than the
                    background concentration determined in Section 8.3.2, whichever
                    concentration would be larger.

             8.3.1.2  If the concentration  of 2,3,7,8-TCDD in the sample is not being checked
                    against a limit specific to that parameter, the spike should be at
                    0.100 l-ig/L or one to five times higher than the background
                    concentration determined in Section 8.3.2,  whichever concentration
                    would be larger.

-------
             8.3.1.3  If it is impractical to determine background levels before spiking (e.g.,
                    maximum holding times will be exceeded), the spike concentration
                    should be (1) the regulatory concentration limit, if any; or, if none
                    (2) the larger of either five times higher than the expected background
                    concentration or 0.100 l-ig/L.

       8.3.2  Analyze one sample aliquot to determine the background concentration (B) of
             2,3,7,8-TCDD.  If necessary, prepare a new QC check sample concentrate
             (Section 8.2.1) appropriate for the background concentration in the sample.
             Spike a second sample aliquot with 1.0 mL of the QC check sample concentrate
             and analyze it to determine the concentration  after spiking (A) of 2,3,7,8-TCDD.
             Calculate percent recovery (P) as  100  (A-B)%T, where T is the known true value
             of the spike.

       8.3.3  Compare the percent recovery (P) for 2,3,7,8-TCDD with the corresponding QC
             acceptance criteria found in Table 2. These acceptance criteria were calculated
             to include an allowance for error  in measurement of both the  background and
             spike concentrations, assuming a spike to background ratio of 5:1. This error
             will be accounted for to the extent that the analyst's spike to background ratio
             approaches 5:1.11  If spiking was performed at a concentration lower than
             0.100 l-ig/L, the analyst must use  either the QC acceptance criteria in Table 2, or
             optional QC acceptance criteria calculated for  the specific spike concentration.
             To calculate optional acceptance criteria for the recovery of 2,3,7,8-TCDD:
             (1) Calculate accuracy (X') using the equation  in Table  3, substituting the spike
             concentration (T) for C; (2) calculate overall precision (S') using the equation  in
             Table 3, substituting X' for X; (3)  calculate the range for recovery at the spike
             concentration as (100 X'/T) ±2.44(100 SVT)%.n

       8.3.4  If the recovery of 2,3,7,8-TCDD falls outside the designated range for recovery,
             a check standard must be analyzed as described in Section 8.4.

8.4     If the recovery of 2,3,7,8-TCDD fails the acceptance criteria for recovery in Section 8.3,
       a QC check standard must be prepared  and analyzed.

       NOTE:    The frequency for the required analysis of a QC check standard will
                 depend upon the complexity of the sample matrix and the performance  of
                 the laboratory.

       8.4.1  Prepare the QC check standard by adding  1.0 mL of QC check sample
             concentrate  (Section 8.2.1 or 8.3.2) to 1 L of reagent water.

       8.4.2  Analyze the QC check standard to determine the concentration measured (A) of
             2,3,7,8-TCDD.  Calculate the percent recovery  (PSK) as  100 (A/T)%, where T is
             the true value of the standard concentration.

       8.4.3  Compare the percent recovery (PSK) with the corresponding QC acceptance
             criteria found in Table 2.  If the recovery of 2,3,7,8-TCDD falls outside the
             designated range, the laboratory performance  is judged to be out of control, and
             the problem must be immediately identified and corrected.  The analytical

-------
             result for 2,3,7,8-TCDD in the unspiked sample is suspect and may not be
             reported for regulatory compliance purposes.

8.5     As part of the QC program for the laboratory, method accuracy for wastewater
       samples must be assessed and records must be maintained. After the analysis of five
       spiked wastewater samples as in Section 8.3, calculate the average percent recovery
       (P) and the spandard deviation of the percent recovery (Sp). Express the accuracy
       assessment as a percent recovery interval from P-2sp to P+2Sp.  If P=90% and^ =10%,
       for example, the accuracy interval is expressed as 70-110%.  Update the accuracy
       assessment on a regular basis (e.g., after each 5-10 new accuracy measurements).

8.6     It is recommended that the laboratory adopt additional quality assurance practices for
       use with this method. The specific practices that are most productive depend upon
       the needs of the laboratory and the nature of the samples.  Field duplicates may be
       analyzed to assess the precision of the environmental measurements.   Whenever
       possible, the laboratory should analyze standard reference materials and participate in
       relevant performance evaluation studies.

9.      Sample Collection, Preservation, and Handling

9.1     Grab samples must be collected in glass containers. Conventional sampling practices12
       should be followed, except that the bottle must not be prerinsed with  sample before
       collection.   Composite samples should be collected in refrigerated glass containers in
       accordance with the requirements of the program.  Automatic sampling equipment
       must be as  free  as possible  of Tygon tubing and other potential sources of
       contamination.

9.2     All samples must be iced or refrigerated at 4°C and protected from light from the
       time of collection until extraction. Fill the sample bottles and, if residual  chlorine is
       present, add 80  mg of sodium thiosulfate per liter of sample and mix well.  EPA
       Methods 330.4 and 330.5  may be used for measurement of residual chlorine.13  Field
       test kits are available for  this purpose.

9.3     Label  all samples and containers "POISON" and ship according to applicable U.S.
       Department of Transportation regulations.

9.4     All samples must be extracted within seven days of collection and completely
       analyzed within 40 days of extraction.2

10.     Sample Extraction

       CAUTION: When  using this method to analyze for 2,3,7,8-TCDD, all of the following
       operations must be performed in a limited-access laboratory with the analyst wearing
       full protective covering for  all exposed  skin surfaces. See Section 4.2.

10.1    Mark  the water meniscus on the side of the sample bottle for later determination of
       sample volume.  Pour the entire sample into a 2 L separatory funnel.

-------
10.2   Add 1.00 mL of internal standard spiking solution to the sample in the separatory
      funnel. If the final extract will be concentrated to a fixed volume below 1.00 mL
      (Section 12.3), only that volume of spiking solution should be added to the sample so
      that the final extract will contain 25 ng/mL  of internal standard at the time of
      analysis.

10.3   Add 60 mL of methylene chloride to the sample bottle, seal, and shake 30 seconds to
      rinse the inner surface.  Transfer the solvent to the separatory funnel and extract the
      sample by shaking the funnel for two minutes with periodic venting to release excess
      pressure. Allow the organic layer to separate from the water phase for a minimum of
      10 minutes.  If the emulsion interface between layers is more than one-third the
      vmlume of the solvent layer, the analyst must employ mechanical techniques to
      complete the phase separation.  The optimum technique  depends upon the sample,
      but may include stirring, filtration of the emulsion through glass wool, centrifugation,
      or other physical methods.  Collect the methylene chloride extract in a 250  mL
      Erlenmeyer flask.

10.4   Add a second 60 mL volume of methylene chloride to the sample bottle and repeat
      the extraction procedure a second time, combining the extracts  in the Erlenmeyer
      flask.  Perform a third extraction in the same manner.

10.5   Assemble a Kuderna-Danish (K-D) concentrator by attaching a  10 mL concentrator
      tube to a 500 mL evaporative flask. Other concentration devices or techniques may be
      used in place of the K-D concentrator if the  requirements of Section 8.2 are met.

10.6   Pour the combined extract into the K-D concentrator. Rinse the Erlenmeyer flask with
      20-30 mL of methylele chloride to complete  the quantitative transfer.

10.7   Add one or two clean boiling chips to the evaporative flask and attach a three-ball
      Snyder column.  Prewet the Snyder column by adding about 1  mL of methylene
      chloride to the top.  Place the K-D apparatus on a hot water bath (60-65°C) so that the
      concentrator tube is partially immersed in the hot water, and the entire lower
      rounded surface of the flask is bathed with  hot vapor.  Adjust the vertical position of
      the apparatus and the water temperature as required to complete the concentration in
      15-20 minutes. At the proper rate of distillation the  balls of the column will actively
      chatter but the chambers will not flood with condensed solvent. When the apparent
      volume of liquid reaches 1 mL, remove the  K-D apparatus and allow it to drain and
      cool for at least  10 minutes.

10.8   Momentarily remove the Snyder column, add 50 mL of hexane and a new boiling
      chip, and reattach the Snyder column. Raise the temperature of the water bath to
      85-90°C.  Concentrate the extract as in Section 10.7, except use hexane to prewet the
      column. Remove the Snyder column and rinse the flask and its lower joint into the
      concentrator tube with 1-2 mL of hexane. A 5 mL syringe  is recommended for this
      operation. Set aside the K-D glassware for reuse in Section 10.14.

10.9   Pour the hexane extract from the concentrator tube into a 125 mL separatory funnel.
      Rinse the concentrator tube  four times with 10 mL aliquots of hexane.  Combine all
      rinses in the 125 mL separatory funnel.

-------
10.10   Add 50 mL of sodium hydroxide solution to the funnel and shake for 30-60 seconds.
       Discard the aqueous phase.

10.11   Perform a second wash of the organic layer with 50 mL of reagent water.  Discard the
       aqueous phase.

10.12   Wash the hexane layer with a least two 50 mL aliquots of concentrated sulfuric acid.
       Continue washing the hexane layer with 50 mL aliquots of concentrated sulfuric acid
       until the acid layer remains colorless. Discard all acid fractions.

10.13   Wash the hexane layer with two 50 mL aliquots of reagent water.  Discard the
       aqueous phases.

10.14   Transfer the hexane extract into a  125 mL Erlenmeyer flask containing 1-2 g of
       anhydrous sodium sulfate.  Swirl the flask for 30 seconds and decant the hexane
       extract into the reassembled K-D apparatus.  Complete the quantitative transfer with
       two 10 mL hexane rinses of the Erlenmeyer flask.

10.15   Replace the one or two clean boiling chips and concentrate the extract to 6-10 mL as
       in Section 10.8.

10.16   Add a clean boiling chip to the concentrator tube and attach a two-ball micro-Snyder
       column.  Prewet the column by adding about 1 mL of hexane to the top.  Place the
       micro-K-D apparatus on the water bath so that the concentrator tube is partially
       immersed in the hot water. Adjust the vertical position of the apparatus and the
       water temperature as required to complete the concentration in 5-10 minutes. At the
       proper rate of distillation the balls of the column will actively chatter but the
       chambers will not flood. When the apparent volume of liquid reaches about 0.5 mL,
       remove the K-D apparatus and allow it to drain and cool for at least 10 minutes.
       Remove the  micro-Snyder  column and rinse its lower joint into the concentrator tube
       with 0.2 mL of hexane.

       Adjust the extract volume  to 1.0 mL with hexane. Stopper the concentrator tube and
       store refrigerated and protected from light if further processing will not be performed
       immediately. If the extract will be stored longer than two days, it should be
       transferred to a Teflon-sealed  screw-cap vial. If the sample  extract requires no further
       cleanup, proceed with GC/MS analysis (Section 12). If the sample requires further
       cleanup, proceed to Section 11.

10.17   Determine the original sample volume by refilling the sample bottle to the mark and
       transferring the liquid to a 1000 mL graduated cylinder.  Record the sample volume to
       the nearest 5 mL.

11.     Cleanup and Separation

11.1    Cleanup procedures may not be necessary for a relatively clean sample matrix.  If
       particular circumstances demand the use of a cleanup procedure, the analyst may use
       either procedure below or  any other appropriate procedure.1'5 7 However, the analyst
       first must demonstrate that the requirements of Section 8.2 can be met using the
       method as revised to incorporate the cleanup procedure.  Two cleanup column

-------
       options are offered to the analyst in this section.  The alumina column should be used
       first to overcome interferences.  If background problems are still encountered, the
       silica gel column may be helpful.

11.2    Alumina column cleanup for 2,3,7,8-TCDD

       11.2.1  Fill a 300 mm long x 10 mm ID chromatographic column with activated
             alumina to the 150 mm level. Tap the column gently to settle the alumina and
             add 10 mm of anhydrous sodium sulfate to the top.

       11.2.2  Preelute the column with 50 mL of hexane. Adjust the elution rate to
             1 mL/min.  Discard the eluate and just prior to exposure of the sodium sulfate
             layer to the air, quantitatively transfer the  1.0 mL sample extract onto the
             column using two 2 mL portions of hexane to complete the transfer.

       11.2.3  Just prior to exposure of the sodium sulfate layer to the air, add  50 mL of 3%
             methylene chloride/95% hexane (V/V) and continue the elution of the column.
             Discard the eluate.

       11.2.4  Next, elute the column with 50 mL of 20% methylene chloride/80% hexane
             (V/V) into a 500  mL K-D flask equipped with a 10 mL concentrator tube.
             Concentrate the collected fraction  to 1.0 mL as in Section  10.16 and analyze by
             GC/MS (Section  12).

11.3    Silica gel column cleanup for 2,3,7,8-TCDD

       11.3.1  Fill a 400 mm long x 11 mm ID chromatmgraphic column with silica  gel to the
             300 mm level.  Tap the column gently to settle the silica gel and  add  10 mm of
             anhydrous sodium sulfate to the top.

       11.3.2  Preelute the column with 50 mL of 20% benzene/80% hexane (V/V).  Adjust
             the elution rate to 1 mL/min. Discard the eluate and just prior to exposure of
             the sodium sulfate layer to the air, quantitatively transfer the  1.0 mL  sample
             extract onto the column using two 2 mL portions of 20% benzene/80% hexane
             to complete the transfer.

       11.3.3  Just prior to exposure of the sodium sulfate layer to the air, add  40 mL of
             20% benzene/80% hexane to the column.  Collect the eluate in a clean 500 mL
             K-D flask equipped with a 10 mL concentrator tube.  Concentrate the collected
             fraction to 1.0 mL as in Section  10.16  and analyze by GC/MS.

12.     GC/MS Analysis

12.1    Table  1 summarizes the recommended operating conditions for  the gas
       chromatograph. Included in this table are retention times and MDL that can be
       achieved under these conditions.  Other capillary columns or chromatographic
       conditions may be used if the requirements of Sections 5.5.2 and 8.2 are met.

-------
12.2    Analyze standards and samples with the mass spectrometer operating in the selected
       ion monitoring (SIM) mode using a dwell time to give at least seven points per peak.
       For LRMS, use masses at m/z 320, 322, and 257 for 2,3,7,8-TCDD and either m/z 328
       for 37C14 2,3,7,8-TCDD or m/z 332 for13Q2  2,3,7,8-TCDD.  For HRMS, use masses at
       m/z 319.8965 and 321.8936 for 2,3,7,8-TCDD and either m/z 327.8847 for 37C14
       2,3,7,8-TCDD or m/z 331.9367 for 13C12 2,3,7,8-TCDD.

12.3    If lower detection limits are required, the extract may be carefully evaporated to
       dryness under a gentle stream of nitrogen with the concentrator tube in a water bath
       at about 40°C. Conduct this operation immediately before GC/MS analysis.
       Redissolve the extract in the desired final volume of ortho-xylene or tetradecane.

12.4    Calibrate the system daily as described in Section 7.

12.5    Inject 2-5 uL of the sample extract into the gas chromatograph.  The volume of
       calibration standard injected must be measured, or be the same as all sample injection
       volumes.

12.6    The presence of 2,3,7,8-TCDD is qualitatively confirmed if all  of the following criteria
       are achieved:

       12.6.1 The gas chromatographic column  must resolve 2,3,7,8-TCDD from the other  21
            TCDD isomers.

       12.6.2 The masses for native 2,3,7,8-TCDD (LRMS-m/z 320, 322, and 257 and
            HRMS-m/z 320 and 322)  and labeled 2,3,7,8-TCDD (m/z 328 or 332) must
            exhibit a simultaneous maximum  at a retention time that matches that of native
            2,3,7,8-TCDD in the calibration standard, with the performance specifications of
            the analytical system.

       12.6.3 The chlorine isotope ratio at m/z  320 and m/z 322 must agree to within ±10%
            of that in the calibration standard.

       12.6.4 The signal of all peaks must be greater than two and one-half times the  noise
            level.

12.7    For quantitation, measure the response of the m/z 320 peak for 2,3,7,8-TCDD and the
       m/z 332 peak for 13C12 2,3,7,8-TCDD or the m/z 328 peak for  37Q1 2,3,7,8-TCDD.

12.8    Co-eluting impurities are suspected if all criteria are achieved except those in
       Section 12.6.3.  In this case, another SIM analysis using masses at m/z 257, 259, 320
       and either m/a 328 or m/z 322  can be performed. The masses at m/z 257 and
       m/z 259 are indicative of the loss of one chlorine and one carbonyl group from
       2,3,7,8-TCDD. If masses m/z 257 and m/z 259 give a chlorine isotope ratio that
       agrees to within ±10% of the same cluster in the  calibration standards, then the
       presence of TCDD  can be confirmed. Co-eluting DDD, DDE,  and PCB residues can be
       confirmed, but will require another injection using the appropriate SIM  masses or full
       repetitive mass scans. If the response for 37C14 2,3,7,8-TCDD at m/z 328 is too  large,
       PCB contamination is suspected and can be confirmed by examining the response at
       both m/z 326 and m/z 328.  The 37C14 2,3,7,8-TCDD internal standard gives negligible

-------
       response at m/z 326. These pesticide residues can be removed using the alumina
       column cleanup procedure.

12.9    If broad background interference restricts the sensitivity of the GC/MS analysis, the
       analyst should employ additional cleanup procedures and reanalyze by GC/MS.

12.10   In those circumstances where these procedures do not yield a definitive conclusion,
       the use of high resolution mass spectrometry is suggested.5

13.     Calculations

13.1    Calculate the concentration of 2,3,7,8-TCDD in the sample using the response factor
       (RF) determined in Section 7.1.2 and Equation 2.

                                      Equation  3"

                                          (As) ds)
                                 RF
                                       (Ais) (RF) (V0)
       where:
             As = Response for the parameter to be measured.
             Ais = Response for the internal standard.
             Is = Amount of internal standard added to each extract (ug).
             V0 = Volume of water extracted (L).

13.2    For each sample, calculate the percent recovery of the internal standard by comparing
       the area of the m/z peak measured in the sample to the area of the same peak in the
       calibration standard.  If the recovery is below 50%, the analyst should review all
       aspects of his analytical technique.

13.3    Report results in l-ig/L without correction for recovery data. All QC data obtained
       should be reported with the sample results.

14.     Method Performance

14.1    The method detection limit (MDL) is defined as the minimum concentration of a
       substance that can be measured and reported with 99% confidence that the value is
       above zero.1  The MDL concentration listed in Table 1 was obtained using reagent
       water.14 The MDL  actually achieved in a given analysis will vary depending on
       instrument sensitivity and matrix effects.

14.2    This method was tested by 11 laboratories using reagent water, drinking water,
       surface water, and three industrial wastewaters spiked at six concentrations over the
       range 0.02-0.20 ug/L.15  Single operator precision, overall precision, and method
       "This equation has been amended to reflect the original as published in 40
CFR 136, FRL-2636-6 49 FR 43234, October 26, 1984.

-------
       accuracy were found to be directly related to the concentration of the parameter and
       essentially independent of the sample matrix.  Linear equations to describe these
       relationships are presented in Table 3.

References

1.      40 CFR Part 136, Appendix B.

2.      "Determination of TCDD in Industrial and Municipal Waste waters,"' EPA
       600/4-82-028, National Technical Information Service, PB82-196882, Springfield,
       Virginia  22161, April 1982.

3.      Buser, H.R. and Rappe, C.  "High Resolution Gas Chromatography of the 22
       Tetrachlorodibenzo-p-dioxin Isomers," Analytical Chemistry, 52, 2257 (1980).

4.      ASTM Annual Book of Standards, Part 31, D3694-78. "Standard Practices for
       Preparation of Sample Containers and for Preservation  of Organic Constituents,"
       American Society for Testing  and Materials,  Philadelphia.

5.      Harless, R. L.,  Oswald, E. O.,  and Wilkinson, M. K.  "Sample Preparation and Gas
       Chromatography/Mass Spectrometry Determination of
       2,3,7,8-Tetrachlorodibenzo-p-dioxin," Analytical Chemistry, 52, 1239 (1980).

6.      Lamparski, L. L. and Nestrick, T. J.  "Determination of  Tetra-, Hepta-, and
       Octachlorodibenzo-p-dioxin Isomers in Particulate Samples at Parts per Trillion
       Levels," Analytical Chemistry,  52, 2045 (1980).

7.      Longhorst, M.  L. and Shadoff, L. A.  "Determination of Parts-per-Trillion
       Concentrations of Tetra-, Hexa-, and Octachlorodibenzo-p-dioxins in Human Milk,"
       Analytical Chemistry, 52, 2037  (1980).

8.      "Carcinogens-Working with Carcinogens," Department of Health, Education, and
       Welfare, Public Health Service, Center for Disease Control, National Institute for
       Occupational Safety and Health, Publication No. 77-206, August 1977.

9.      "OSHA Safety and Health Standards, General  Industry," (29 CFR Part 1910),
       Occuptional Safety and Health Administration, OSHA 2206 (Revised, January 1976).

10.     "Safety in Academic Chemistry Laboratories,"  American Chemical Society Publication,
       Committee on Chemical Safety, 3rd Edition,  1979.

11.     Provost, L. P. and Elder, R. S.  "Interpretation  of Percent Recovery Data," American
       Laboratory, 15,  58-63 (1983). (The value 2.44 used  in the equation in Section 8.3.3 is
       two times the value 1.22 derived in this report.)

12.     ASTM Annual Book of Standards, Part 31, D3370-76, "Standard Practices for Sampling
       Water," American Society for Testing and Materials, Philadelphia.

-------
13.     "Methods, 330.4 (Titrimetric, DPD-FAS) and 330.5 (Spectrophotometric DPD) for
       Chlorine, Total Residual," Methods for Chemical Analysis of Water and Wastes,
       EPA-600/4-79-020, U.S. Environmental Protection Agency, Environmental Monitoring
       and Support Laboratory, Cincinnati, Ohio 45268, March 1979.

14.     Wong, A.S. et al. "The Determination of 2,3,7,8-TCDD in Industrial and Municipal
       Wastewaters, Method 613, Part 1-Development and Detection Limits," G. Choudhay,
       L. Keith, and C. Ruppe, ed., Butterworth Inc.,  (1983).

15.     "EPA Method Study 26, Method 613: 2,3,7,8-Tetrachlorodibenzo-p-dioxin," EPA
       600/4-84-037, National Technical Information Service, PB84-188879, Springfield,
       Virginia  22161, May 1984.


           Table 1—Chromatographic Conditions and Method Detection Limit
Parameter
2,3,7,8-TCDD 	

Retension
time (min)
13.1

Method detection
limit (ug/L)
0.002

Column conditions:  SP-2330 coated on a 60 m long x 0.25 mm ID glass column with
hydrogen carrier gas at 40 cm/sec, linear velocity, splitless injection using tetradecane.
Column temperature held isothermal at 200°C for one minute, then programmed at
8°C/min. to 250°C and held. Use of helium carrier gas will approximately double the
retention time.
                     Table 2—QC Acceptance Criteria—Method 613
Parameter
2,3,7,8-TCDD 	

Test
Cone.
(ug/L)
0.100

Limit for s
(ug/L)
0.0276

Range for X
(ug/L)
0.0523-0.1226

Range for
P, Ps (%)
45-129

s_= Standard deviation of four recovery measurements, in u/L (Section 8.2.4).
X = Average recovery for four recovery measurements, in u/L (Section 8.2.4).
P, Ps = Percent recovery measured (Section 8.3.2, Section 8.4.2).

NOTE:      These criteria are based directly upon the method performance data in Table 3.
            Where necessary, the limits for recovery have been broadened to assure
            applicability of the limits to concentrations below those used to develop
            Table 3.

-------
  Table 3—Method Accuracy and Precision as Functions of Concentration—Method 613
Parameter
2-Chloronaphthalene

Accuracy, as
recovery,
X' (ug/L)
075C+321

Single analyst,
precision,
sr,(ug/L)
0280X-1 17

Overall
precision,
S' (p-g/L)
038X-1 39

X' = Expected recovery for one or more measurements, of a sample containing a
concentration of C, in l-ig/L.
sr' = Expected single analyst standard deviation of measurements at an average concentration
found of X.8, in l-ig/L.
S' = Expected interlaboratory standard deviation of measurements at an average
concentration found of X, in l-ig/L.
C = True value for the concentration, in l-ig/L.
X = Average recovery found for measurements of samples containing a concentration of C,
in l-ig/L.

-------