SEPA
United States
Environmental Protection
Agency
 www.epa.gov
  Promulgated 1984
   Method 612: Chlorinated
   Hydrocarbons

-------
                             APPENDIX A TO PART 136
       METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND
                            INDUSTRIAL WASTEWATER

                 METHOD 612—CHLORINATED HYDROCARBONS
1.

1.1
Scope and Application

This method covers the determination of certain chlorinated hydrocarbons. The
following parameters can be determined by this method:
Parameter
2-Chloronaphthalene 	
1,2-Dichlorobenzene 	
1,3-Dichlorobenzen 	
1,4-Dichlorobenzene 	
Hexachlorobenzene 	
Hexachlorobutadiene 	
Hexachlorocyclopentadiene 	
Hexachloroethane 	
1,2,4-Trichlorobenzene 	

STORET No.
34581
34536
34566
34571
39700
34391
34386
34396
34551

CAS No.
91-58-7
95-50-1
541-73-1
106-46-7
118-74-1
87-68-3
77-47-4
67-72-1
120-82-1

1.2
1.3
1.4
This is a gas chromatographic (GC) method applicable to the determination of the
compounds listed above in municipal and industrial discharges as provided under
40 CFR Part 136.1.  When this method is used to analyze unfamiliar samples for any
or all of the compounds above, compound identifications should be supported by at
least one additional qualitative technique.  This method describes a second gas
chromatographic column that can be  used to confirm measurements made with the
primary column. Method 625 provides gas chromatograph/mass spectrometer
(GC/MS) conditions appropriate for the qualitative and quantitative confirmation of
results for all of the parameters listed above, using the extract produced by this
method.

The method detection limit (MDL, defined in Section 14.1)1 for each parameter is
listed in Table 1. The MDL for a specific  wastewater may differ from those listed,
depending upon the nature of interferences in the  sample matrix.

The sample extraction and concentration steps in this method are essentially the same
as in Methods 606, 608, 609, and 611. Thus, a single sample may be extracted to
measure the parameters included  in the scope of each of these methods. When
cleanup is required, the concentration levels must  be high enough to permit selecting
aliquots, as necessary, to apply appropriate cleanup procedures.  The analyst is
allowed the latitude, under Section 12, to  select chromatographic conditions
appropriate for the simultaneous measurement of  combinations of these parameters.

-------
1.5     Any modification of this method, beyond those expressly permitted, shall be
       considered as a major modification subject to application and approval of alternate
       test procedures under 40 CFR Parts 136.4 and 136.5.

1.6     This method is restricted to use by or under the supervision of analysts experienced
       in the use of a gas chromatograph and in the interpretation of gas chromatograms.
       Each analyst must demonstrate the ability to generate acceptable results with this
       method using the procedure described in Section 8.2.

2.      Summary of Method

2.1     A measured volume of sample, approximately 1 L, is extracted with methylene
       chloride using a separatory funnel.  The methylene chloride extract is dried and
       exchanged to  hexane during concentration to a volume of 10 mL or less. The extract
       is separated by gas chromatography and the parameters are then measured with an
       electron capture detector.2

2.2     The method provides  a Florisil column cleanup procedure to aid in the elimination of
       interferences that may be encountered.

3.      Interferences

3.1     Method interferences may be caused by contaminants in solvents, reagents, glassware,
       and other sample processing hardware that lead to discrete artifacts and/or elevated
       baselines in gas chromatograms.  All of these materials must be routinely
       demonstrated to be  free from interferences under the conditions of the analysis by
       running laboratory reagent blanks as described in Section 8.1.3.

       3.1.1   Glassware must be scrupulously  cleaned.3  Clean all glassware as soon as
              possible after use by rinsing with the last solvent used in it.  Solvent rinsing
              should be  followed by detergent washing with hot water, and rinses with tap
              water and distilled water. The glassware should then be drained dry, and
              heated  in a muffle furnace at 400°C for  15-30 minutes. Some thermally stable
              materials, such as PCBs, may not be eliminated by this treatment.  Solvent
              rinses with acetone and pesticide quality hexane may be substituted for the
              muffle furnace heating.  Thorough rinsing with such solvents usually
              eliminates PCB interference.  Volumetric ware should not be heated in a muffle
              furnace. After drying and cooling, glassware should be sealed  and stored in a
              clean environment to prevent any accumulation of dust or other contaminants.
              Store inverted  or capped with aluminum foil.

       3.1.2   The use of high purity reagents and solvents helps to minimize interference
              problems.  Purification of solvents by distillation in all-glass systems may be
              required.

3.2     Matrix interferences may be caused by contaminants that are co-extracted from the
       sample. The extent of matrix interferences will vary considerably from source to
       source, depending upon the nature and diversity of the industrial complex or
       municipality being sampled. The cleanup procedure in Section  11 can be used to

-------
       overcome many of these interferences, but unique samples may require additional
       cleanup approaches to achieve the MDL listed in Table 1.

4.      Safety

4.1     The toxicity or carcinogenicity of each reagent used in this method has not been
       precisely defined; however, each chemical compound should be treated as a potential
       health hazard.  From this viewpoint, exposure to these chemicals must be reduced to
       the lowest possible level by whatever means available. The  laboratory is responsible
       for maintaining a current awareness file of OSHA regulations regarding the safe
       handling of the chemicals specified  in this method.  A reference file of material data
       handling sheets should also be made available to all personnel involved in the
       chemical analysis.  Additional references to laboratory safety are available and have
       been identified4 6 for the information of the analyst.

5.      Apparatus and Materials

5.1     Sampling equipment, for discrete  or composite sampling.

       5.1.1    Grab sample bottle—1 L or 1 qt,  amber glass, fitted with a screw cap lined
              with Teflon. Foil may be substituted for Teflon if the sample is not corrosive.
              If amber bottles are not available, protect samples from light. The bottle and
              cap liner must be washed, rinsed with acetone or methylene chloride, and
              dried before use to minimize contamination.

       5.1.2    Automatic sampler (optional)—The sampler  must incorporate glass sample
              containers for the collection of a minimum of 250 mL of sample.  Sample
              containers must be kept refrigerated at 4°C and protected from light during
              compositing.  If the sampler  uses a peristaltic pump,  a minimum length of
              compressible silicone rubber tubing may be used.  Before use, however, the
              compressible tubing should be thoroughly rinsed with methanol, followed by
              repeated rinsings with distilled water to minimize the potential for
              contamination of the sample. An integrating flow meter is required to  collect
              flow proportional composites.

5.2     Glassware (All specifications are suggested.  Catalog numbers are included for
       illustration  only.)

       5.2.1    Separatory funnel—2 L, with Teflon stopcock.

       5.2.2    Drying column—Chromatographic column, approximately 400 mm long x
              19 mm ID, with coarse frit filter disc.

       5.2.3    Chromatographic column—300 long x 10 mm ID, with Teflon stopcock and
              coarse frit filter disc at bottom.

       5.2.4    Concentrator tube, Kuderna-Danish—10 mL, graduated (Kontes K-570050-1025
              or equivalent). Calibration must be checked at the volumes employed in the
              test.  Ground glass stopper is used to prevent evaporation of extracts.

-------
       5.2.5  Evaporative flask, Kuderna-Danish— 500 mL (Kontes K-570001-0500 or
             equivalent).  Attach to concentrator tube with springs.

       5.2.6  Snyder column, Kuderna-Danish—Three-ball macro (Kontes K-503000-0121 or
             equivalent).

       5.2.7  Vials—10-15 mL, amber glass, with Teflon-lined screw cap.

5.3     Boiling chips—Approximately 10/40 mesh. Heat to 400°C for 30  minutes or Soxhlet
       extract with methylene chloride.

5.4     Water bath—Heated, with concentric ring cover, capable of temperature control
       (±2°C).  The bath should be used in a hood.

5.5     Balance—Analytical, capable of accurately weighing 0.0001 g.

5.6     Gas chromatograph—An analytical system complete with gas chromatograph suitable
       for on-column injection and all required accessories including syringes, analytical
       columns, gases, detector, and strip-chart recorder.  A data system is recommended for
       measuring peak areas.

       5.6.1  Column 1—1.8 m long x 2 mm ID glass, packed with 1% SP-1000 on
             Supelcoport (100/120 mesh) or equivalent.  Guidelines for the use of alternate
             column packings are provide in Section 12.1.

       5.6.2  Column 2—1.8 m long x 2 mm ID glass, packed with 1.5% OV-1/2.4% OV-225
             on Supelcoport (80/100 mesh) or equivalent. This column was used to
             develop the method performance statements in Section 14.

       5.6.3  Detector—Electron capture detector. This detector has proven effective in the
             analysis of wastewaters for the parameters listed in the scope (Section 1.1), and
             was used to develop the method performance statements in Section 14.
             Guidelines for the use of alternate detectors are provided in Section 12.1.

6.      Reagents

6.1     Reagent water—Reagent water is defined as a water in which an interferent is not
       observed at the MDL of the parameters of interest.

6.2     Acetone, hexane,  isooctane, methanol, methylene chloride, petroleum ether (boiling
       range 30-60°C)—Pesticide quality or equivalent.

6.3     Sodium  sulfate—(ACS) Granular, anhydrous.  Purify heating at 400°C for four hours in
       a shallow tray.

6.4     Florisil—PR grade (60/100 mesh).  Purchase activated at 1250°F and store in the  dark
       in glass  containers with ground glass stoppers or foil-lined screw caps.  Before use,
       activate  each batch at least  16 hours at 130°C  in a foil-covered glass container and
       allow to cool.

-------
6.5     Stock standard solution (1.00 ug/uL)—Stock standard solutions can be prepared from
       pure standard materials or purchased as certified solutions.

       6.5.1   Prepare stock standard solutions by accurately weighing about 0.0100 g of
              pure material.  Dissolve the material in isooctane and dilute to volume in a
              120 mL volumetric flask. Larger volumes can be used at the convenience of
              the analyst.  When compound purity is assayed to be 96% or greater, the
              weight can be used without correction to calculate the concentration of the
              stock standard.  Commercially prepared stock standards can be used at any
              concentration if they are certified by the manufacturer or by an independent
              source.

       6.5.2   Transfer the  stock standard solutions into Teflon-sealed screw-cap bottles.
              Store at 4°C  and protect from light. Stock standard solutions should be
              checked frequently for signs of degradation or evaporation, especially just
              prior to preparing calibration standards from them.

       6.5.3   Stock standard solutions must be replaced after six months, or sooner if
              comparision with check standards  indicates a problem.

6.6     Quality control check sample concentrate—See Section 8.2.1.

7.      Calibration

7.1     Establish gas chromatographic operating conditions equivalent to those given in
       Table 1.  The gas chromatographic system can be calibrated  using the external
       standard technique  (Section 7.2) or the internal standard technique (Section 7.3).

7.2     External standard calibration procedure

       7.2.1   Prepare calibration standards at a minimum of three  concentration levels for
              each parameter of interest by adding volumes of one or more stock standards
              to a volumetric flask and diluting to volume with isooctane. One of the
              external standards should be at a concentration near, but above, the MDL
              (Table  1)  and the other concentrations should correspond to the expected range
              of concentrations found in real samples or should define the working range of
              the detector.

       7.2.2   Using injections of 2-5 uL, analyze each calibration standard according to
              Section 12 and tabulate peak height or area responses against the  mass
              injected.  The results can be used to prepare a calibration curve for each
              compound.  Alternatively, if the  ratio of response to amount injected
              (calibration factor) is a constant over the working range (<10% relative
              standard deviation,  RSD),  linearity through the origin can be assumed and  the
              average ratio or calibration factor can be used in place of a calibration curve.

7.3     Internal standard calibration procedure—To use  this approach, the analyst must select
       one or more internal standards that are  similar in analytical  behavior  to the
       compounds of interest.  The analyst must  further demonstrate that the measurement
       of the internal standard is not affected by  method or matrix  interferences. Because of

-------
       these limitations, no internal standard can be suggested that is applicable to all
       samples.

       7.3.1   Prepare calibration standards at a minimum of three concentration levels for
             each parameter of interest by adding volumes of one or more stock standards
             to a volumetric flask. To each calibration standard, add a known constant
             amount of one or more internal standards, and dilute to volume with
             isooctane.  One of the standards should be at a concentration near, but above,
             the MDL and the other concentrations should correspond to the expected
             range of concentrations found in real samples or should define the working
             range of the detector.

       7.3.2   Using injections of 2-5 uL, analyze each calibration standard according to
             Section 12  and tabulate peak height or area responses against concentration for
             each compound and internal standard.  Calculate response factors (RF) for
             each compound using Equation 1.

                                       Equation 1


                                             (Cis)
                                   RF =
                                         (Ais) (Cs)
             where:
                    As = Response for the parameter to be measured.
                    Ais = Response for the internal standard.
                    Cis = Concentration of the internal standard (ug/L).
                    Cs = Concentration of the parameter to be measured (ug/L).

             If the RF value over the working range is a constant (<10% RSD), the RF can
             be assumed to be invariant and the average RF can be used for calculations.
             Alternatively, the results can be  used to plot a calibration curve of response
             ratios, As/Ais, vs. concentration ratios Q /Qs *.

7.4     The working calibration curve, calibration factor, or RF must be verified on each
       working day by the measurement of one or more calibration standards. If the
       response for any parameter varies from the predicted response by more than ±15%, a
       new calibration curve must be prepared for that compound.

7.5     Before using any cleanup procedure, the analyst must process a series of calibration
       standards through the procedure to validate elution patterns and the absence of
       interferences from the reagents.
       "This equation corrects an error made in the original method publication (49 FR
43234, October 26, 1984). This correction will be formalized through a rulemaking in FY97.

-------
8.      Quality Control

8.1     Each laboratory that uses this method is required to operate a formal quality control
       program.  The minimum requirements of this program consist of an initial
       demonstration of laboratory capability and an ongoing analysis of spiked samples to
       evaluate and document data quality.  The  laboratory must maintain records to
       document the quality of data that is generated.  Ongoing data quality checks are
       compared with established performance criteria to determine if the results of analyses
       meet the performance characteristics of the method.  When the results of sample
       spikes indicate atypical method performance, a quality control check standard must be
       analyzed to confirm that the measurements were performed in an in-control mode of
       operation.

       8.1.1  The analyst must make an initial, one-time, demonstration of the ability to
             generate acceptable accuracy and precision with this method. This ability is
             established as described in Section  8.2.

       8.1.2  In recognition of advances that are occurring in chromatography, the analyst is
             permitted certain options (detailed  in Sections 10.4, 11.1, and 12.1) to improve
             the separations or lower the cost of measurements.  Each time such
             modification is made to the method, the analyst is required  to repeat the
             procedure in Section 8.2.

       8.1.3  Before processing any samples, the analyst must analyze a reagent water blank
             to demonstrate that interferences from the analytical system and glassware are
             under control.  Each time a set of samples is extracted or reagents are changed,
             a reagent water blank must be processed as a safeguard against laboratory
             contamination.

       8.1.4  The laboratory must, on an ongoing basis, spike and analyze a minimum of
             10% of all samples to monitor and  evaluate laboratory data  quality.  This
             procedure is described in Section 8.3.

       8.1.5  The laboratory must, on an ongoing basis, demonstrate through the analyses of
             quality control check standards that the operation of the measurement system
             is in control. This procedure is described in Section 8.4. The frequency of the
             check standard analyses is equivalent to 10% of all samples  analyzed but may
             be reduced if spike recoveries from samples (Section 8.3) meet all specified
             quality control criteria.

       8.1.6  The laboratory must maintain performance records to document the quality of
             data that is generated. This procedure is described in Section 8.5.

8.2     To establish  the ability to generate acceptable accuracy and precision, the analyst must
       perform the  following operations.

       8.2.1  A quality control (QC) check sample concentrate is required containing each
             parameter of interest at the following concentrations in acetone:
             Hexachloro-substituted parameters, 10 uKg/mL; any other chlorinated
             hydrocarbon, 100 uKg/mL. The QC check sample concentrate must be

-------
              obtained from the U.S. Environmental Protection Agency, Environmental
              Monitoring and Support Laboratory in Cincinnati, Ohio, if available. If not
              available from that source, the QC check sample concentrate must be obtained
              from another external source.  If not available from either source above, the
              QC check sample concentrate must be prepared by the laboratory using stock
              standards prepared independently from those used for calibration.

       8.2.2   Using a pipet, prepare QC check samples at the test concentrations shown in
              Table 2 by adding  1.00 mL of QC check sample concentrate to each of four 1 L
              aliquots of reagent water.

       8.2.3   Analyze the well-mixed QC check samples according to the method beginning
              in Section 10.

       8.2.4   Calculate the average recovery (X) in uKg/L, and the standard deviation of
              the recovery (s)  in uKg/L, for each parameter using the four results.

       8.2.5   For each parameter compare s and X  with the corresponding acceptance  _
              criteria for precision and accuracy, respectively, found in Table 2. If s and X
              for all parameters of  interest meet the acceptance criteria, the system
              performance is acceptable and analysis of actual samples can begin. If any
              individual s exceeds the precision limit or  any individual X  falls outside the
              range for accuracy, the system performance is unacceptable for that parameter.

              NOTE:       The large number of parameters in Table 2 presents a substantial
                           probability that one or more will fail at least one of the
                           acceptance criteria when all parameters are analyzed.

       8.2.6   When one or more of the parameters tested fail at least one of the acceptance
              criteria, the analyst must proceed according to Section 8.2.6.1 or 8.2.6.2.

              8.2.6.1 Locate and correct the source of the problem and repeat the test for all
                    parameters  of interest beginning with Section 8.2.2.

              8.2.6.2 Beginning with Section 8.2.2, repeat the test only for those parameters
                    that failed to meet criteria.  Repeated failure, however, will confirm a
                    general problem with the measurement system. If this occurs, locate
                    and correct the source of the problem and repeat the  test for all
                    compounds of interest beginning with Section 8.2.2.

8.3     The laboratory must, on an ongoing basis, spike at least 10% of the  samples from each
       sample site being monitored to assess accuracy. For laboratories analyzing one to ten
       samples per month, at least one spike sample per month is required.

       8.3.1   The concentration of  the spike in the sample should be determined as follows:

              8.3.1.1 If, as in compliance monitoring, the concentration of a specific
                    parameter in the sample is being checked against a regulatory
                    concentration  limit, the spike should be at that limit or one to five times

-------
                    higher than the background concentration determined in Section 8.3.2,
                    whichever concentration would be larger.

              8.3.1.2 If the concentration of a specific parameter in the sample is not being
                    checked against a limit specific to that parameter, the spike should be
                    at the test concentration in Section 8.2.2 or one to five times higher than
                    the background concentration determined in Section 8.3.2,  whichever
                    concentration would be larger.

              8.3.1.3 If it is impractical to determine background levels before spiking (e.g.,
                    maximum holding times will be exceeded), the spike concentration
                    should  be (1)  the regulatory concentration limit, if any; or, if none by
                    (2)  the larger  of either five times higher than the expected  background
                    concentration or the test concentration in Section 8.2.2.

       8.3.2   Analyze one sample  aliquot to determine the background concentration (B) of
              each parameter.  In necessary, prepare a new QC check sample concentrate
              (Section  8.2.1) appropriate for the background concentrations in the sample.
              Spike a second sample aliquot with 1.0 mL of the QC check sample concentrate
              and analyze it  to determine the concentration after spiking (A) of each
              parameter. Calculate each percent recovery (P) as 100 (A-B)%/T,  where T is
              the known true value of the spike.

       8.3.3   Compare the percent recovery (P) for each parameter with the corresponding
              QC  acceptance criteria found  in Table 2.  These acceptance criteria were
              calculated to include an allowance for error in measurement of both the
              background and spike concentrations, assuming a spike to background ratio of
              5:1.  This error will be accounted for to the extent that the analyst's spike to
              background ratio approaches  5:1.7 If spiking was performed at a concentration
              lower than the test concentration in Section 8.2.2, the analyst must use either
              the QC acceptance criteria in Table 2, or optional QC acceptance criteria
              calculated for the specific spike concentration. To calculate optional acceptance
              criteria for the  recovery of a parameter:  (1) Calculate accuracy (X') using the
              equation in Table 3, substituting the spike concentration (T) for C; (2)  calculate
              overall precision (S')  using the equation in Table 3, substituting X' for X;
              (3) calculate the range for recovery at the spike concentration as (100 X'/T)
              ±2.44 (100 SVT)%.7

       8.3.4   If any individual P falls outside the designated range for recovery, that
              parameter has  failed the acceptance criteria.  A check standard containing each
              parameter that failed the criteria must be analyzed as described in Section 8.4.

8.4.    If any parameter fails  the acceptance criteria for recovery in Section 8.3, a QC check
       standard containing each parameter that failed must be prepared and analyzed.

       NOTE:       The frequency for the required analysis of a QC check standard will
                    depend upon the number of parameters being simultaneously  tested,
                    the complexity of the sample matrix, and the performance of the
                    laboratory.

-------
       8.4.1  Prepare the QC check standard by adding 1.0 mL of QC check sample
             concentrate (Sections 8.2.1 or 8.3.2) to 1 L of reagent water. The QC check
             standard needs only to contain the parameters that failed criteria in the test in
             Section 8.3.

       8.4.2  Analyze the QC check standard to determine the concentration measured (A)
             of each parameter.  Calculate each percent recovery (PJ as 100 (A/T)%, where
             T is the true value of the standard concentration.

       8.4.3  Compare the percent recovery (PJ for each parameter with the corresponding
             QC acceptance criteria found in Table 2.  Only parameters that failed the test
             in Section 8.3 need to be compared with  these criteria.  If the recovery of any
             such parameter falls outside the designated range, the laboratory performance
             for that parameter is judged to be out of control, and the problem must be
             immediately identified and corrected. The analytical result for that parameter
             in the unspiked sample is suspect and may not be reported for regulatory
             compliance purposes.

8.5     As part  of the QC program for the laboratory, method accuracy for wastewater
       samples must be assessed and records must  be maintained. After the analysis of five
       spiked wastewater samples as in Section 8.3, calculate the average percent recovery
       (P) and the standard deviation of the percent recovery (Sp).  Express the accuracy
       assessment as a percent recovery interval from P-2sp to  P+2Sp.  If P=90% and^ =10%,
       for example, the accuracy interval is expressed as 70-110%.  Update the accuracy
       assessment for each parameter  on a regular basis (e.g., after each 5-10 new accuracy
       measurements).

8.6     It is recommended that the  laboratory adopt additional quality assurance practices for
       use with this method.  The specific practices that are most productive depend upon
       the needs of the laboratory and the nature of the samples.  Field duplicates may be
       analyzed to assess the precision of the environmental measurements. When doubt
       exists over the identification of a peak on the chromatogram, confirmatory techniques
       such as gas chromatography with a dissimilar column, specific element detector, or
       mass spectrometer must be used. Whenever possible, the laboratory should analyze
       standard reference materials and participate  in relevent performance evaluation
       studies.

9.      Sample  Collection, Preservation, and Handling

9.1     Grab samples must be collected in glass  containers.  Conventional sampling practices8
       should be followed, except that the bottle must not be prerinsed with sample before
       collection.  Composite samples should be collected in refrigerated glass containers  in
       accordance with the requirements of the program. Automatic sampling equipment
       must be as free as possible of Tygon tubing and other potential sources of
       contamination.

9.2     All samples must be iced or refrigerated at 4°C from the time of collection until
       extraction.

-------
9.3     All samples must be extracted within seven days of collection and completely
       analyzed within 40 days of extraction.2

10.     Sample Extraction

10.1    Mark the water meniscus on the side of the sample bottle for later determination of
       sample volume.  Pour the entire sample into a 2 L separatory funnel.

10.2    Add 60 mL of methylele chloride to the sample bottle, seal, and shake 30 seconds to
       rinse the inner surface. Transfer the solvent to the separatory funnel and extract the
       sample by shaking the funnel for two  minutes with periodic venting to release excess
       pressure. Allow the organic layer to separate from the water phase for a minimum of
       10 minutes.  If the emulsion interface between layers is more than one-third the
       volume of the solvent layer, the analyst must employ mechanical techniques to
       complete the phase separation.  The optimum technique  depends upon the sample,
       but may include stirring, filtration of the emulsion through glass wool, centrifugation,
       or other physical methods. Collect the methylene chloride extract in a 250 mL
       Erlenmeyer flask.

10.3    Add a second 60 mL volume  of methylene chloride to the sample bottle and repeat
       the extraction procedure a second time, combining the extracts in the Erlenmeyer
       flask. Perform a third extraction in the same manner.

10.4    Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10 mL  concentrator
       tube  to a 500  mL evaporative flask.  Other concentration devices or techniques may be
       used in place of the K-D concentrator  if the requirements of Section 8.2 are met.

10.5    Pour the combined extract through a solvent-rinsed drying  column containing about
       10 cm of anhydrous sodium sulfate, and collect the extract in the K-D concentrator.
       Rinse the Erlenmeyer flask and  column with 20-30 mL of methylene chloride to
       complete the quantitative transfer.

10.6    Add one or two clean boiling chips to the evaporative flask and attach a three-ball
       Snyder column.  Prewet the Snyder column by adding about 1 mL  of methylene
       chloride to the top.  Place the K-D apparatus on a hot water bath (60-65°C) so that the
       concentrator tube is partially immersed in the hot water, and the entire lower
       rounded surface of the flask is bathed with hot vapor. Adjust the vertical position  of
       the apparatus and the water temperature as required to complete the concentration in
       15-20 minutes.  At the proper rate of distillation the balls of the column will actively
       chatter but the chambers will not flood with condensed solvent. When the apparent
       volume of liquid reaches 1-2 mL, remove the K-D apparatus and allow it to drain and
       cool for at least 10 minutes.

       NOTE:       The dichloribenzenes have a sufficiently high volatility that significant
                    losses may occur  in concentration steps if care is not exercised.  It is
                    important to maintain a constant gentle evaporation rate and not to
                    allow the liquid volume to fall below 1-2 mL before removing the K-D
                    apparatus from the hot  water bath.

-------
10.7    Momentarily remove the Snyder column, add 50 mL of hexane and a new boiling
       chip, and reattach the Snyder column. Raise the tempeature of the water bath to
       85-90°C.  Concentrate the extract as in Section 10.6, except use hexane to prewet the
       column. The elapsed time of concentration should be 5-10 minutes.

10.8    Romove the Snyder column and rinse the flask and its lower joint into the
       concentrator tube with  1-2 mL of hexane. A 5 mL syringe is recommended for this
       operation. Stopper the  concentrator tube and store refrigerated if further processing
       will not be performed immediately.  If the extract will be stored longer than two days,
       it should be transferred to a Teflon-sealed screw-cap vial. If the sample extract
       requires no further cleanup, proceed with gas chromatographic analysis (Section  12).
       If the sample requires further cleanup, proceed to Section 11.

10.9    Determine the original  sample volume by refilling the sample bottle to the mark and
       transferring the liquid to a 1000 mL graduated cylinder.  Record the sample volume to
       the nearest  5 mL.

11.     Cleanup and Separation

11.1    Cleanup procedures may not be necessary for a relatively clean sample matrix. If
       particular circumstances demand the use of a cleanup procedure, the analyst may use
       the procedure below or any other appropriate procedure. However, the analyst first
       must demonstrate that  the requirements of Section  8.2  can be met using the method as
       revised to incorporate the cleanup procedure.

11.2    Florisil column cleanup for chlorinated hydrocarbons

       11.2.1  Adjust the sample extract to 10 mL with hexane.

       11.2.2  Place 12  g of Florisil into a chromatographic column.  Tap the column  to settle
             the Florisil and add 1-2 cm of anhydrous sodium sulfate to the top.

       11.2.3  Preelute  the column with 100 mL of petroleum ether. Discard the eluate and
             just prior to exposure of the sodium sulfate  layer to the air, quantitatively
             transfer the sample extract onto the column by  decantation and subsequent
             petroleum ether washings. Discard the eluate.  Just prior to exposure of the
             sodium sulfate layer to the air, begin eluting the column with 200 mL of
             petroleum ether and collect the eluate in a 500 mL K-D flask  equipped with a
              10 mL concentrator tube. This fraction should contain all  of the chlorinated
             hydrocarbons.

       11.2.4  Concentrate the  fraction as in Section 10.6, except use hexane to prewet the
             column.  When  the apparatus is cool, remove the Snyder column and rinse the
             flask and its lower joint into the concentrator tube with hexane.  Analyze by
             gas chromatography (Section 12).

-------
12.     Gas Chromatography

12.1    Table 1 summarizes the recommended operating conditions for the gas
       chromatograph. Included in this table are retention times and MDL that can be
       achieved under these conditions. Examples of the separations achieved by Columl 2
       are shown in Figures 1 and 2. Other packed or capillary (open-tubular) columns,
       chromatographic conditions, or detectors may be used if the requirements  of
       Section 8.2 are  met.

12.2    Calibrate the system daily as described in Section 7.

12.3    If the internal standard calibration procedure is being used, the internal standard
       must be added to the sample extract and mixed throughly immediately before
       injection into the gas chromatograph.

12.4    Inject 2-5 uL of the sample extract or standard into the gas chromatograph using the
       solvent-flush techlique.9  Smaller (1.0 uL) volumes may be injected if automatic
       devices are employed. Record the volume injected to the nearest 0.05 uL,  the total
       extract volume, and the resulting peak size in area or peak height units.

12.5    Identify the parameters in the sample by comparing the retention times of the peaks
       in the sample chromatogram with those of the peaks in standard chromatograms.
       The width of the retention time window used to make identifications should be based
       upon measurements of actual retention time variations of standards over the course of
       a day.  Three times the standard deviation of a retention time for a compound can be
       used to calculate a suggested window  size; however, the experience of the analyst
       should  weigh heavily in  the interpretation of chromatograms.

12.6    If the response for a peak exceeds the working range of the system, dilute  the extract
       and reanalyze.

12.7    If the measurement of the peak response is prevented by the presence of interferences,
       further  cleanup is required.

13.     Calculations

13.1    Determine the concentration of individual compounds in the sample.

       13.1.1  If the external standard calibration procedure is used, calculate the  amount of
             material injected from the peak response using the calibration curve or
             calibration factor  determined in Section 7.2.2.  The concentration in the sample
             can be calculated  from Equation 2.

-------
                                       Equation 2

                                                  (A) (V.)
                            Concentration (ug/L) =
             where:
                    A = Amount of material injected (ng).
                    V; = Volume of extract injected (uL) .
                    Vt = Volume of total extract (uL) .
                    Vs = Volume of water extracted (mL) .

       13.1.2 If the internal standard calibration procedure is used, calculate the
             concentration in the sample using the response factor (RF) determined in
             Section 7.3.2  and Equation 3.

                                       Equation 3

                                                  (As) (Is)
                         Concentration (ug/L)  =
                                               (Ajs) (RF) (V0)


             where:
                    As = Response for the parameter to be measured.
                    Ais = Response for the internal standard.
                    Is = Amount of internal standard added to each extract (ug).
                    V0 = Volume of water extracted (L).

13.2    Report results in l-ig/L without correction for  recovery data.  All QC data obtained
       should be reported with the sample results.

14.     Method Performance

14.1    The method detection limit (MDL) is defined as the minimum concentration of a
       substance that can be measured and reported with 99% confidence that the value is
       above zero.1 The MDL concentrations listed in Table 1 were obtained using reagent
       water.10  Similar results were achieved using representative wastewaters.  The MDL
       actually  achieved in a given analysis will vary depending on instrument sensitivity
       and matrix effects.

14.2    This method has been tested for linearity of spike recovery from reagent water and
       has been demonstrated to  be applicable over the concentration range from 4 x MDL to
       1000 x MDL.10

14.3    This method was tested by 20 laboratories using reagent water, drinking water,
       surface water, and three industrial wastewaters spiked at six concentrations over the
       range  1.0-356 ug/L.11  Single operator precision, overall precision, and method
       accuracy were found to be directly related to the concentration of the parameter and
       essentially independent of the sample matrix.  Linear equations to describe these
       relationships are presented in Table 3.

-------
References

1.      40 CFR Part 136, Appendix B.

2.      "Determination of Chlorinated Hydrocarbons In Industrial and Municipal
       Wastewaters," EPA 6090/4-84-ABC, National Technical Information Service, PBXYZ,
       Springfield, Virginia  22161 November 1984.

3.      ASTM Annual Book of Standards, Part 31, D3694-78.  "Standard Practices for
       Preparation of Sample Containers and for Preservation of Organic Constituents,"
       American Society for Testing and Materials, Philadelphia.

4.      "Carcinogens-Working With Carcinogens," Department of Health, Education, and
       Welfare, Public Health Service, Center for Disease Control, National Institute for
       Occupational Safety and Health, Publication No. 77-206, August 1977.

5.      "OSHA Safety and Health Standards, General Industry,"  (29 CFR Part 1910),
       Occupational Safety and Health Administration, OSHA 2206 (Revised, January 1976).

6.      "Safety in Academic Chemistry Laboratories," American  Chemical Society Publication,
       Committee on Chemical Safety, 3rd Edition, 1979.

7.      Provost, L.P. and Elder, R.S.  "Interpretation of Percent Recovery Data,"American
       Laboratory, 15, 58-63 (1983).  (The  value 2.44 used in the equation in Section 8.3.3 is
       two times the value 1.22 derived  in this report.)

8.      ASTM Annual Book of Standards, Part 31, D3370-76.  "Standard Practices for
       Sampling Water," American Society for Testing and Materials, Philadelphia.

9.      Burke, J.A.  "Gas Chromatography for Pesticide Residue  Analysis; Some Practical
       Aspects," Journal of the Association of Official Analytical Chemists, 48, 1037 (1965).

10.     "Development of Detection Limits,  EPA Method 612, Chlorinated Hydrocarbons,"
       Special letter report for EPA Contract 68-03-2625, U.S. Environmental Protection
       Agency, Environmental Monitoring and Support Laboratory, Cincinnati, Ohio  45268.

11.     "EPA Method Study Method 612-Chlorinated Hydrocarbons,"  EPA 600/4-84-039,
       National Technical Information Service, PB84-187772, Springfield, Virginia  22161,
       May 1984.

12.     "Method Performance for Hexachlorocyclopentadiene by Method 612," Memorandum
       from R. Slater, U.S. Environmental  Protection Agency, Environmental Monitoring  and
       Support Laboratory, Cincinnati, Ohio 45268, December 7, 1983.

-------
          Table 1—Chromatographic Conditions and Method Detection Limits
Parameter

1 3-Dichlorobenzene
Hexchloroethane
1 4-Dichlorobenzene
1 2-Dichlorobenzene
Hexachlorobutadiene
1 2 4-Trichlorobenzene
Hexachlorocyclopentadiene
2-Chloronaphthalene 	
Hexachlorobenzene

Retention
Column 1
45
49
52
66
77
15 5
nd
a2.7
a56

:ime (min)
Column 2
68
83
76
93
200
223
C165
b3.6
b!0 1

Method
detection
limit (ug/L)
1 19
003
1 34
1 14
034
005
040
0.94
005

Column 1 conditions:  Supelcoport (100/120 mesh) coated with 1% SP-1000 packed in a
1.8 m x 2 mm ID glass column with 5% methane/95% argon carrier gas at 25 mL/min. flow
rate. Column temperature held isothermal at 65°C, except where otherwise indicated.

Column 2 conditions:  Supelcoport (80/100 mesh) coated with 1.5% OV-1/2.4% OV-225
packed in a 1.8 m x 2 mm ID glass column with 5% methane/95% argon carrier gas at
25 mL/min. flow rate. Column temperature held isothermal at 75°C, except where
otherwise indicated.

nd = Not determined.
a!50°C column temperature.
b!65°C column temperature.
C100°C column temperature.

-------
                     Table 2—QC Acceptance Criteria—Method 612
Parameter
2-Chloronaphthalene
1 2-Dichlorobenzene
1 3-Dichlorobenzene
1 4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobuladiene
Hexachlorocyclopentadiene
Hexachloroethane
1 2 4-Trichlorobenzene

Test Cone.
(ug/L)
100
100
100
100
10
10
10
10
100

Limit for s
(ug/L)
373
283
264
208
24
2 2
2 5
33
31 6

Range for X
(ug/L)
29 5-1269
23 5-145 1
72-1386
22 7-1269
2 6-148
D-12 7
D-104
2 4-123
20 2-1337

Range for
P, Ps (%)
9-148
9-160
D-150
13-137
15-159
D-139
D-lll
8-139
5-149

s_= Standard deviation of four recovery measurements, in l-ig/L (Section 8.2.4).
X = Average recovery for four recovery measurements, in l-ig/L (Section 8.2.4).
P, Ps = Percent recovery measured (Section 8.3.2, Section 8.4.2).
D = Detected; result must be greater than zero.

NOTE:       These criteria are based directly upon the method performance data in Table 3.
             Where necessary, the limits for recovery have been broadened to assure
             applicability  of the limits to  concentrations below those used to develop
             Table 3.
   Table 3—Method Accuracy and Precision as Functions of Concentration—Method 612
Parameter
2-Chloronaphthalene 	
1 2-Dichlorobenzene
1 3-Dichlorobenzene
1 4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobuladiene
Hexachlorocyclopentadiene
Hexachloroethane
1 2 4-Trichlorobenzene

Accuracy, as
recovery, X'
(ug/L)
0.75C+3.21
0 85C-0 70
0 72C+0 87
0 72C+2 80
0 87C-0 02
061C+003
047C
0 74C-0 02
0 76C+0 98

Single analyst
precision, S/
(ug/L)
0.280X-1.17
0 22X-2 95
0 21X-1 03
0 16X-048
0 14X+007
0 18X+008
024X
023X+007
023X-044

Overall
precision, S'
(ug/L)
0.38X-1.39
041X-392
049X-398
035X-057
036X-0 19
053X-0 12
050X
036X-000
040X-1 37

X = Expected recovery tor one or more measurements ot a sample containing a
concentration of C, in l-ig/L.
sr' = Expected single analyst standard deviation of measurements at an average concentration
found of X, in l-ig/L.
S' = Expected interlaboratory standard deviation of measurements at an average
concentration found of X, in l-ig/L.
C = True value for the concentration, in l-ig/L.
X = Average recovery found for measurements of samples containing a concentration of C,
in l-ig/L.
Estimates based upon the performance in a single laboratory.

-------
COLUMN: 1.5% OV-1/2.4% OV-225 ON SUPELCOPORT
TEMPERATURE:  75%
DETECTOR:  aECTRON CAPTURE
              8     12    16     20    24

                 RETENTION TIME. MIN.
   Figure 1. Gas chromatogram of chlorinated hydrocarbons.

-------
COLUMN: 1.5% OV-1/2.4% OV-225 ON SUPELCOPORT

TEMPERATURE: 165*C

DETECTOR: ELECTRON CAPTURE
                II
                   o
                    I
                   CM
                             IS)
                             u
                             <
               0    4    8   12


               RETENTION TIME, MIN.
   Figure 2. Gas chromatogram of chlorinated hydrocarbons.

-------