United States                          EPA Documents 740-S1-5001
       Environmental Protection                   September 2015
       Agency                             Office of Chemical Safety and
                                        Pollution Prevention
Indoor Exposure Product Testing Protocols
                    Version  1.0
                     September 2015

-------
Table of Contents

Introduction	4
Exposure Testing Protocol 1: Source Characterization	6
Exposure Testing Protocol 2: Short-Term Emission Testing	8
Exposure Testing Protocol 3: Long-Term Emissions from Articles - Partition and Diffusion Coefficients. 15
Exposure Testing Protocol 4: Particulate Matter Formation Due to Mechanical Forces Applied to Product
           or Article Surfaces	23
Exposure Testing Protocol 5: Photolysis under Simulated Indoor Lighting Conditions	27
Exposure Testing Protocol 6: Oral Exposure - Migration Rate and Transfer Efficiency	32
Exposure Testing Protocol 7: Dermal Exposure - Potential Exposure	37


Tables

Table 1.    Commonly used environmental chambers for testing of chemical emissions from products
and articles3	8
Table 2.    Methods for experimental determination of partition and diffusion coefficients	15


Figures

Figure 1.    Conceptual  Diagram of source-to-dose continuum for consumer products and articles	5
Figure 2.    Schematic of example 30 m3 full-scale chamber (Liu et al., 2012)	9
Figure 3.    Schematic diagram of small-scale VOC emission chamber (Yerramilli et al., 2010)	10
Figure 4.    Schematic plot of Field  and Laboratory Emission Cell (FLEC) (Kim et al., 2007) a) horizontal
view; b) schematic view	10
Figure 5.    Photo of micro-chamber/ thermal extractor (u.CTE) from Markes International, Llantrisant,
UK (Cleanroom Technology, 2011)	10
Figure 6.    Schematic plot of the microbalance test system (Cox et al., 2001)	17
Figure 7.    Schematic plot of the dynamic-static chamber (He et al., 2010)	17
Figure 8.    Schematic plot of the diffusionmetric method (Bodalal et al., 2000)	18
Figure 9    Schematic plot of the dual-chamber method (Xiong et al., 2009)	19
Figure 10.    Schematic plot of the dual chamber method (Liu et al., 2014)	19
Figure 11.    Schematic plot of the cup method (Blondeau et al., 2003)	20
Figure 12.    Schematic of the test facility for particle generation due to abrasion	23
Figure 13.    Graphic example of generic procedure for photolysis without dust	29
Figure 14.    Graphic example of generic procedure for photolysis with dust	30
Figure 15.    Graphic example of procedure for analyzing migration from product or article surface to
saliva        	34
Figure 16.    Graphic example of small-scale procedure for analyzing migration from product or article
surface to sweat	40
                                         Page 2 of 41

-------
AUTHORS, CONTRIBUTORS and ACKNOWLEDGEMENTS
This report was developed by the United States Environmental Protection Agency (US EPA),
Office of Chemical Safety and Pollution Prevention (OCSPP), Office of Pollution Prevention and
Toxics (OPPT). Mention of trade names does not constitute endorsement by the EPA.

EPA Authors:
Charles Bevington, OPPT/RAD
Christina Cinalli, OPPT/RAD (retired)
Cathy Fehrenbacher, OPPT/RAD
Zhishi Guo, ORD (retired)

Acknowledgements
The following individuals contributed to portions of this document:
Xiaoyu Liu, ORD
Treye Thomas, CPSC
Kathleen Ernst, NIOSH (retired)

This document was developed with support from ICF International under EPA Contract # EP-W-
12-010.
                                  Page 3 of 41

-------
Introduction

EPA's Office of Pollution Prevention and Toxics has developed a set of seven indoor exposure product
testing protocols intended to provide information on the purpose of the testing, general description of
the sampling and analytical procedures, and references for a base-set of exposure tests that will be used
to inform and refine estimates of indoor exposures. The scope of these protocols is limited to testing
chemicals in products or articles/building used in indoor environments. These protocols are general in
nature and will need to be tailored to the specific type of chemical to be analyzed, the particular product
or article which is being evaluated and the data quality objective for the testing.

The  protocols are intended to be used in combination to evaluate  potential exposures when using
products and articles in indoor environments. For example if the testing objective is to evaluate how much
of a particular chemical is emitted during a short-term use of a particular product indoors, the source
characterization protocol and the short-term emission test protocol would be appropriate. The protocols
would be modified to  include the appropriate analytical method for the chemical of  interest, the
appropriate type of chamber, sample preparation, sampling method, sampling volume, etc.

The  protocols should be modified using methodologies generally accepted in the  relevant scientific
community at the time the study is initiated. Before starting to conduct any study that will use a modified
version of these protocols, a written test protocol is generally submitted to the Agency for review. During
the Agency review of the modified protocol, a review of the data quality objective, the sampling process
design (experimental  design), sampling  and analytical methods, sample handling and custody, quality
control procedures and activities (including reference samples, duplicates, replicates, etc.), instruments
and equipment to  be used in conducting the testing, data review, verification, and validation, as well as
reporting requirements. Additional  information on  the Agency's Quality Analysis procedures  and
programs is available (EPA. 2011). The final report shall contain study results and sufficient contextualizing
information on testing conditions and analytical approaches to inform study results.

 Each study shall  be conducted in good faith, with due care, and  in a scientifically valid manner. The
protocols are listed below; they may be updated overtime:
        Source Characterization
concentration (ppm), weight fraction (0-1)
        Short-Term Emission Testing
emission rate (mass/time)
emission factor (mass/area/time)
        Long-Term Emission Testing
solid-phase diffusion coefficient (Dm)
material-air partition coefficient (Kma)
        Particulate Matter Formation Due to Mechanical
        Forces Applied to Product/Article Surfaces
particle generation rate (mass/time)
        Photolysis under Simulated Indoor
        Lighting Conditions
time-averaged air, wipe, and/or dust concentrations
        Oral Exposure: Mouthing of Objects
        and Transfer Efficiency
migration rate mass/surface area/time
transfer efficiency (fraction)
        Dermal Exposure: Potential Exposure
mass/surface area/event on skin

-------
Figure 1 provides an illustration of the types of potential exposures associated with the source (product
or article in  the indoor environment), and how the exposure data produced from  applying the test
protocols will be used to inform the potential for exposure.
             Protocol #5
             Photolysis
Inhalation and Oral Dose
    from Inhaled Air
                        Migration of Chemical
                        from Product Surface
                         throughout Indoor
                            Environment
                                                       Protocol #6
                                                    Mouthing Exposure
             Protocol #2
          Short-Term Emission
           Protocol #3
        Long-Term Emission
             ProtocolfH
           Paniculate Matter
             Formation
                                                                 Protocol 87
                                                               Dermal Exposure
            Protocol 81
       Source Characterization
                                                                         RECEPTOR
Figure 1.    Conceptual Diagram of source-to-dose continuum for consumer products and articles.
                                            Page 5 of 41

-------
Exposure Testing Protocol 1:  Source Characterization

Purpose:
To collect basic information on the properties associated with the behavior of the chemical when it is used
within various end-use applications.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario of interest.

Description:
The exposure potential of a chemical  used in an  article or product is influenced by several parameters.
Chemicals that are part of formulated mixtures are generally liquids or semi-solids and are used overtime
and disposed. Chemicals that are added  to  articles or building materials are generally part of solid
matrices. The likelihood of a chemical migrating from an article is dependent on the characteristics of the
material of which the article is comprised as well as the chemical itself.

For example, polyurethane foam produced for specific purposes may have varying specifications for
properties such as density, rigidity, and structure (closed vs. open cell)  along with the thickness of the
product and its exposed surface area. These properties influence the likelihood of migration and are thus
important in  understanding the potential  for exposure. The overall impact of one or a combination of
these factors that could influence migration and exposure potential is not well characterized.

The objective of the Source Characterization  protocol is to determine the concentration (ppm)  and/or
weight fraction  (0-1) of the chemical present within the article, building material, or consumer product.
Additional contextualizing information that may be required  (depending on the specific chemical and
product tested)  includes:

    Physical-chemical properties that govern the behavior of the chemical in the indoor environment,
    including: Henry's Law constant, octanol-water partitioning coefficient, octanol-air partitioning
    coefficient, water solubility, and vapor pressure. Properties may need to be measured or adjusted
    for relevant indoor environment and/or body temperatures reflecting conditions of use. Expected
    temperatures during use should be reported.
    Information characterizing the type and properties of the material. Properties of the material
    include density, rigidity, porosity, surface area, and thickness.
    Information characterizing the properties of the product. Properties of the consumer product
    include density, physical form, method of application, and whether dilution occurs during routine
    use.
    Use category descriptions including clear and specific definitions.
    The typical setting for use (e.g., outdoors,  indoors,  residential, commercial).
    Typical life expectancy of the article during use, typical or high-end mass of product used per event,
    and duration of use per event.
                                          Page 6 of 41

-------
Reporting of Results and Records Retention:
A final report shall be  prepared, and records  shall  be retained  in accordance  with  40 CFR 792,
Subpart J - Records and Reports.

For example, the following types of key information should be included in the report:

   Sampling and analytical methods — description or citation, including deviation from standard
   procedure, if applicable.
   Quality Assurance/Quality Control data: accuracy and precision of measurements

References:
Product Properties Test Guidelines OPPTS 830.1550. Product Identity and Composition, available at:
   http://www.regulations.gov/#!documentDetail;D=EPA-HQ-OPPT-2009-0151-0003
OPPT Voluntary Use and Exposure Information Project Form, available at:
   http://www.epa.gov/oppt/exposure/pubs/ueipform.pdf
                                         Page 7 of 41

-------
Exposure Testing Protocol 2: Short-Term Emission Testing

Purpose:
To collect information on emission rates of chemicals from products or articles through chamber testing.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario of interest.

Description:

Approach
Chemical emissions from products and articles are most commonly tested in environmental chambers,
which are designed based on the theory of continuous stirred tank reactor (CSTR) in chemical engineering.
Thus,  many principals  of the CSTR are  applicable to test chambers, including mixing, residence time,
steady state, and the assumption that the chemical concentration in the outlet air is representative of
that inside the chamber. Atypical chamber system consists of the chamber itself, clean air supply, airflow
control, air sampling ports, temperature and humidity sensors and controls, and data acquisition system.
An electric fan is  often installed in small and  large chambers, known as the conventional chambers, to
improve air mixing and maintain certain  air speed. Typical test conditions are 23°C, 50% relative humidity
and 0.1 m/s air speed. The air change rate varies depending on chamber types. Over time, progress has
been made to standardize testing for certain kinds of materials in certain kinds of chambers. The major
chamber types are summarized  in Table 1. More standard methods have been, or are being, developed
for testing specific chemicals/materials — such as California Department of Public Health/Environmental
Health Laboratory Branch (CDPH/EHLB) standard method for California Specification 01350 (2010), ASTM
D6007, ASTM WK40293, ANSI/BIFMA M7.1  and  ANSI/BIFMA x7.1 —  but they are all based  on  the
standards in Table 1.

Table 1.   Commonly used environmental chambers for testing of chemical emissions from
          products and articles3
Chamber
Type
Full-scale
chamber
Small-scale
chamber
Micro
chamber
Field and Laboratory
Emission Cells
Typical
Size
30m3
SOL
0.05-0. 1L
0.035 L
Typical Air
Change Rate (Iv1)
1
1
>100
>100
Commercially
Available
No
Yes
Yes
Yes
References
ISO 16000-9
ASTM D6670
ISO 16000-9
ASTM D5116
ISO 12219-3
ASTM D7706
ISO 16000-10
ASTM D7143
aMid-scale chambers, typically 1 to 10 m3 in size, are also available but less commonly used.
                                         Page 8 of 41

-------
Selection of Test Chambers
Selecting a chamber suitable for testing a given chemical in a given product or article depends on several
factors, including the properties of the chemical of interest and those of the substrate. A general guideline
is provided below.

The full-scale chamber is most suitable for testing VOC emissions from article assemblies such as furniture,
computers, TV sets, portable air cleaning devices, home electronics, and office equipment. The full-scale
chamber is more costly to operate than other types of chambers and can accommodate testing of large
items.
            Equipment
              room       Low efficiency  Air
                      ,—K,  air filer  conditioner
                      Activated alumina
                      impregnated with
    Low efficiency Activated   potassium  High efficiency
Blower   air filter  carbon filter Permanganate    air filter
                                                            Low efficiency  Air
                                                              air filter  conditioner
            Test room
                          VOC   rS
                       sampling tube
Figure 2.   Schematic of example 30 m3 full-scale chamber (Liu et al., 2012)

The small-scale chamber is suitable for volatile organic compound (VOC) emissions from a large variety of
products and articles as long as they can be cut into coupons or panels that fit the chamber size. It has
limited capability for testing semi-volatile organic compound (SVOC) emissions.
               Clean Air
               Humidity
               control
               system
                                                 PiiT.-r

                                               Chamber*!

                                              T«stSp*cim«n
                                           Inlet    f    Outlet

D                                               Chambers!

                                               TestSpecim en
                                         Temperature Controlled Cabinet
                                            Page 9 of 41

-------
Figure 3.   Schematic diagram of small-scale VOC emission chamber (Yerramilli et al., 2010)

The Field and Laboratory Emission Cell (FLEC) has a cone-shaped cavity and can be placed directly on the
surface of the test material, which becomes the bottom of the cell. Because of its small volume (35 ml)
and large source area (20 cm in diameter), FLEC has the largest loading factor among all test chambers. It
is mostly used to test VOC emissions from building materials with a flat and non-porous surface. It has
limited capability for testing SVOC emissions.
                 Silicon rubber foam
                                                                                Air outlet
                                             Sealing material

                                                       Slit
Figure 4.   Schematic plot of Field and Laboratory Emission Cell (FLEC) (Kim et al., 2007)
           a) horizontal view; b) schematic view

Micro  chambers are small  cells operated  at  a  high air
exchange  rate.  These chambers are suitable for rapid
screen-ing of material emissions and have been used for
both VOCs and SVOCs.
Temperature  and  humidity  controls are important for
emissions testing. While all of these chambers can meet
these requirements, the  micro  chambers have  a  wider
range  of  temperature  control  and, thus, are  more
convenient for testing emissions at elevated temperature.
Figure 5.
Photo of micro-chamber/
thermal extractor (uCTE)
from Markes International,
Llantrisant, UK (Cleanroom
Technology, 2011)
Testing of SVOC emissions is more challenging than testing
VOCs because the interior surfaces of the test chamber can
adsorb  a significant  amount of  SVOCs  from  air.  In
conventional test chambers and FLEC, most SVOCs emitted from the source are adsorbed by interior
surfaces (Clausen et al., 2004). This problem can be somewhat alleviated by using micro chambers, which
have a high air change rate and relatively small surface area. An alternative is to use a specially-designed
chamber that is  modified to minimize the sink effect (Xu et al., 2012).

When the SVOC emissions cannot be detected at room temperature, testing at elevated temperatures
can be considered. In order to extrapolate the test results to normal temperature, tests should be
conducted at a minimum of three temperatures.
                                        Page 10 of 41

-------
Sample Preparation, Transport, Storage, and Conditioning
Most standards,  including those shown  in Table  1,  contain  a section  for  sample  preparation,
transportation, storage, and conditioning. The California standard method (CDPH/EHLB, 2010) contains
more details about this subject. There are also stand-alone standards for sample handling (e.g., ISO 16000-
11). The main goal is to prevent the samples from being contaminated or losing representativeness due
to exposure to extreme conditions such as contaminated air or materials, light, excessive moisture, and
elevated temperature.

To prepare test specimens, flat products/articles are cut into coupons (or panels). The size of the test
specimen is often expressed as a loading factor (the exposed surface area divided by the volume of the
test chamber, in (m2/m3). For the same test specimen, a large loading factor means higher concentrations
in chamber air.

Generic Test Procedure
    Prior to a test, clean the chamber according to the procedure in the standard methods for the
    chamber.
    Check the chamber for leakage.
    Flush the chamber with clean air at the specified air flow rate, temperature, and humidity; take  a
    background air sample to ensure that the chamber is free of contamination.
    Open the lid (or door) of the chamber to place the test specimen(s) into the chamber (Note that in
    conventional chambers, the test specimen is often placed in the center of the chamber floor. To
    increase the loading factor, test specimens can also be placed vertically by using a rack).
    Close and tighten the chamber lid (or door) and record the test start time.
    Collect air samples according the sampling plan (see below for more details).

The test duration depends on the source type and data needs. To determine  emission trends (constant
versus  decaying emissions), a minimum of one week is recommended, during which at least one half
dozen samples should be taken at different elapsed time.

To calculate emission rate or emission factor for non-constant sources, more samples (e.g., a dozen) are
often needed. Because the chamber concentration changes rapidly in the early hours of testing, higher
sampling frequency is needed in  the early hours. This  is especially  important for conventional test
chambers.

Sampling Methods
Selection of the sampling method requires consideration of several factors, including collection efficiency,
specificity,  capacity  (potential breakthrough), and compatibility with the analytical  methods. Many
general-purpose sampling methods have been developed for collecting VOCs and SVOCs from chamber
air, including sorbent tubes (Tenax, XAD  resins, charcoal,  silica gel etc.),  impingers,  filters, and
polyurethane foam  samplers. There are  also chemical specific  sampling  media.  For  example, 2,4
dinitrophenylhydrazine cartridges  are commonly used for sampling aldehydes  (ASTM D6803).

Sampling Volume
Whether the  chemicals of interest in the emissions can be captured in air samples  depends on the
sensitivity of the analytical methods and sample volume. A low sample volume may result in no detection
of the chemical of interest. Thus, it is important to roughly determine the proper sampling volume before
testing starts. This is often achieved in two ways: (1) trial and error, which is done by conducting a pilot
or scouting test; and (2) estimating the order-of-magnitude of the air concentration based on existing
mass transfer models, from which a proper sample  volume can be determined when  the  method
                                         Page 11 of 41

-------
quantification limit is known. This method requires knowledge of mass transfer source  models and
parameter estimation methods, however.

Sample Analysis
Many standard methods can be used to analyze the air samples collected from chamber testing (e.g., EPA
Methods TO-01, TO-17, 3545A and 8270D; ASTM D 7339 and D 5197; ISO 16000-3 and ISO 16000-6). Gas
chromatography  (GC)  with different  detectors  (e.g.,  flame ionization, electron capture, and  mass
spectrometry detectors) are most commonly used for VOC and SVOC analysis. High performance liquid
chromatography (HPLC) is often used for aldehydes and some SVOCs. Commonly used detectors include
UV, fluorescence, and tandem mass spectrometry.

Reporting Results and Records  Retention:
A final report shall be  prepared, and  records  shall  be retained in accordance with 40 CFR 792,
Subpart J - Records and Reports.

The standard test methods mentioned above contain sections for reporting, which may aid  in preparing
the report of results. For example, the following key information should be included in the report:

    Test article: article name, manufacture and/or purchase date, origin, intended use, uniformity
    (homogeneous, layered, spray application, coating, etc.), dimensions of test specimens, density,
    exposed area, treatment of sample edges (sealed or exposed) and information about sample
    creation, transport, and storage.
    Target chemical(s) and their basic properties: CAS number, molecular formula, vapor pressure,
    chemical reactivity, content/percent within the material, etc.
    Test chamber: chamber type, volume, loading, dimensions, and interior surface material (e.g.,
    polished stainless steel, PTFE-coated stainless steel,  Silicosteel-coated stainless steel, and glass).
    Test procedure: description or citation, including deviation from standard procedure.
    Sampling and analytical methods: description or citation, including deviation from standard
    procedure. Description of accuracy and precision.
    Environmental conditions: chamber air flow rates, temperatures, relative humidity values, and air
    exchange rates expressed in arithmetic mean and standard deviation.
    Test results: concentration vs. time. ASTM 5116 describes a method to convert chamber
    concentrations to emission rate (in mass/time) and emission factor (in mass/area/time).
    QA/QC data: accuracy and precision of measurements, calibrations, daily calibration checks,
    background samples, blank samples.

References:
ANSI/BIFMA M7.1-2011: Standard Test Method for determining VOC Emissions from Office Furniture
    Systems, Components and Seating, available at:
    https://www.bifma.org/store/ViewProduct.aspx?id=1375383
ANSI/BIFMA X7.1-2011 Standard for Formaldehyde and TVOC Emissions of Low-emitting Office Furniture
    and Seating, available at https://www.bifma.org/store/ViewProduct.aspx?id=1375803
ASTM D5116-10 Standard Guide for Small-Scale Environmental Chamber Determinations of Organic
    Emissions from Indoor Materials/Products, available at
    http://compass.astm.org/EDIT/html annot.cgi?D5116+10
ASTM D6007-14 Standard Test Method for Determining  Formaldehyde Concentrations in Air from Wood
    Products Using a Small-Scale Chamber, available at http://www.astm.org/Standards/D6007.htm
                                        Page 12 of 41

-------
ASTM 6670-01 (2007) Standard Practice for Full-Scale Chamber Determination of Volatile Organic
    Emissions from Indoor Materials/Products, available at http://www.astm.org/Standards/D6670.htm
ASTM D6803-13 Standard Practice for Testing and Sampling of Volatile Organic Compounds (Including
    Carbonyl Compounds) Emitted from Paint Using Small Environmental Chambers, available at
    http://www.astm.org/Standards/D6803.htm
ASTM D7143 -11 Standard Practice for Emission Cells for the Determination of Volatile Organic
    Emissions from Indoor Materials/Products, available at http://www.astm.org/Standards/D7143.htm
ASTM D7706-11 Standard Practice for Rapid Screening of VOC Emissions from Products Using Micro-
    Scale Chambers,  available at http://www.astm.org/Standards/D7706.htm
ASTM E1333-10 Standard Method for Determining Formaldehyde Concentrations in Air and Emission
    Rates from Wood Products Using a Large Chamber, available at
    http://www.astm. org/search/fullsite-search.html?querv=E1333&
ASTM WK40293 New test method for estimating chemical emissions from spray polyurethane foam
    (SPF) insulation using micro-scale environmental test chambers, available at
    http://www.astm.org/DATABASE.CART/WORKITEMS/WK40293.htm
CDPH/EHLB (2010). Standard Method VI. 1, Standard method for the testing and evaluation of volatile
    organic chemical emissions from indoor sources using environmental chambers. Version 1.1..
    available at
    https://www.scsglobalservices.com/files/standards/CDPH  EHLB StandardMethod VI  1 2010.pdf
Clausen, P., Hansen, V., Gunnarsen, L, Afshari, A., Wolkoff, P. (2004). Emission of di-2-ethylhexyl
    phthalate from PVC flooring into air and uptake in dust: emission and sorption experiments in FLEC
    and CLIMPAQ. Environmental Science & Technologies, 38,  2531-2537.
Cleanroom Technology (2011). VOC emissions test method, available at
    http://www.cleanroomtechnology.com/
    technical/article  page/VOC emissions  testULmethod/58849
EPA Method 8260B Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS)
EPA Method 8270D: Semi volatile Organic Compounds by Gas Chromatography/Mass Spectrometry
    (GC/MS), available at http://www3.epa.gov/epawaste/hazard/testmethods/sw846/pdfs/8270d.pdf
EPA (2011). Quality Management Tools - QA Project Plans. http://www.epa.gov/QUALITY/qapps.html
ISO 16000-6:2011 Indoor air - Part 6: Determination of volatile organic compounds in indoor and test
    chamber air by active sampling on Tenax TA sorbent, thermal desorption and gas chromatography
    using MS or MS-FID
ISO 16000-9: Indoor Air-Part 9: Determination of the Emission of Volatile Organic Compounds from
    Building Products and Furnishing-Emission Test Chamber Method, available at
    http://www.iso.org/iso/catalogue detail.htm?csnumber=38203
ISO (2006). ISO 16000-10:2006 - Indoor air - Part 10: Determination of the emission of volatile organic
    compounds from building products and furnishing - Emission test cell method, available at
    http://www.iso.org/iso/iso catalogue/catalogue tc/catalogue detail.htm?csnumber=38204
Kim S, Kim J, and Kim H (2007). Application of Field and Laboratory Emission Cell (FLEC) to Determine
    Formaldehyde and VOCs Emissions from Wood-Based Composites. Mokchae Konghak 35(5): 24-37.
Liu W, Zhang Y, and Yao Y (2012). Labeling of volatile organic compounds emissions from Chinese
    furniture: Consideration and practice. Chinese Science Bulletin, 58: 3499-3506
                                        Page 13 of 41

-------
Xu Y, Liu Z, Park J, Clausen PA, Banning JL, and Little JC (2012). Measuring and predicting the emission
    rate of phthalate plasticizer from vinyl flooring in a specially-designed chamber. Environmental
    Science and Technology, 46: 12534-12541.
Yerramilli S, Schiller R, Downie R, and Garnys V (2010). Measurement of Chemical Emissions from
    Building Products. The Australian Building Services Journal, 1: 41-44.
                                         Page 14 of 41

-------
Exposure Testing Protocol 3: Long-Term Emissions from Articles -
Partition and Diffusion Coefficients

Purpose:
To collect information on physical/chemical properties that influence migration rates of volatile and semi-
volatile organic compounds (VOCs and SVOCs) into the indoor environment.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario  of interest.

Description:

Basics of partition and diffusion coefficients
Volatile and semi-volatile organic compounds (VOCs and SVOCs) emitted from solid surfaces (e.g., building
materials, consumer products) can affect indoor air quality (Cox et al., 2001). Because testing long-term
emissions is costly and time-consuming, mass transfer  models have been developed to predict the
emission and transport of chemicals. Initial concentration in the source (C0), the solid-phase diffusion
coefficient (Dm), material-air partition coefficient (Kma), and gas-phase mass transfer coefficient (h) are key
parameters that impact the emissions. For new  products and articles, C0 can be estimated based on
product formulation and parameter  h is often estimated with empirical models. Therefore, the partition
and diffusion coefficients are  key to understanding the long-term effect of chemical emission from
products and articles.

Theoretically, the  diffusion transport of molecules is related to the properties of the chemical such as
molecular weight, molecular size (volume or area), and the molecular polarity; the properties of the
substrate; and environmental conditions such as temperature,  air velocity, and  relative humidity. The
material-air partition coefficient is often correlated with the volatility of the chemical and properties of
the substrate. While there are many methods for experimental determination of Dm and  Kma, standard
methods are lacking. No single method is suitable for testing all materials  and chemicals.  Most existing
methods are suitable for VOCs only.

Methods to estimate partition and diffusion coefficients
Table 2 summarizes eight experimental methods for measuring the partition and diffusion coefficients for
solid materials. Details associated with each method are described below.

Table 2.    Methods for experimental determination of partition and diffusion coefficients
         Method
       Microbalance
Yes
Yes
                        Applicability
VOCs
                                  Reference
                                                                          Cox et al., 2001
                                                                         Zhao et al., 2004


No
Yes
Yes
Yes
VOCs
VOCs
Meininghausetal., 2002
He et al., 2010
                                        Page 15 of 41

-------
   Static diffusion metric
         method
Yes
Yes
VOCs
  Bodalaletal., 2000
Table 2.    Methods for experimental determination of partition and diffusion coefficients
           (continued)
         Method
   Twin dynamic chamber
         methods
Yes
Yes
                         Applicability
VOCs
                                   Reference
   Xiong et al., 2009;
    Xuetal., 2012
Meininghausetal., 2000
Meininghausetal., 2002
Dual chamber in series
Variable volume loading
Cup method
Porosity-based method
Yes
Yes
No
No
Yes
No
Yes
Yes
SVOCs
VOCs
VOCs
VOCs
Liu et al., 2014
Xiong etal., 2011
Kirchneretal., 1999
Blondeau et al., 2003
Microbalance method
The microbalance method can be used to estimate the partition and diffusion coefficients by placing the
test specimen on a microbalance located in a dynamic chamber with temperature and humidity control,
as shown in Figure 6 (Cox et al., 2001; Zhao et al., 2004). In the beginning of the test, the sample weight
is first stabilized by passing clean air through the chamber until an equilibrium is obtained. The sorption
process begins by introducing an air stream with a constant and known concentration  of VOC into the
chamber. The mass gain of the test specimen due to VOC sorption overtime is monitored. The monitoring
continues for a period of time  after the equilibrium is reached. During the desorption process, the
chamber is purged with clean air and the weight loss of the test specimen is monitored until an equilibrium
is re-established. This is a gravimetric method. With the sorption and  desorption data measured by the
microbalance, the partition coefficient is determine by the ratio of the solid- and gas-phase concentrations
and the diffusion coefficient by non-linear regression.
                                         Page 16 of 41

-------
                             Constant Temperature Enclosure
                                  Temperature
                                 Transducer (RTD)
Figure 6.   Schematic plot of the microbalance test system (Cox et al., 2001)

Dynamic-static chamber method
The system of the dynamic-static chamber method is composed of a Field and Laboratory Emission Cell
(FLEC), a static chamber (test chamber), and a measurement device.  One example of a measurement
device is a  proton transfer reaction-mass spectrometer (PTR-MS)  (Meininghaus et al., 2002; He et al.,
2010). The static chamber serves as a limited reservoir for gaseous VOCs. The test material, as a thin plate
with uniform thickness, is placed between the FLEC and the static chamber. During the test, clean gas
(VOC free) from a compressed air cylinder passes through the FLEC at a controlled rate (Figure 7), and
VOC is introduced to the static chamber at a  certain concentration. The VOC in the static chamber will
diffuse to the FLEC through the  test material driven by the concentration gradient. The real-time VOC
concentration in the outlet air of the FLEC  is sampled and analyzed by an appropriate method. The
concentration data is used to estimate the partition and diffusion coefficients.
                            Mass flow controller
                                              FLEC
Figure 7.   Schematic plot of the dynamic-static chamber (He et al., 2010)
                                         Page 17 of 41

-------
Static diffusion metric method
The diffusion metric method uses a twin static diffusion chamber system to determine the diffusion
coefficient (Bodalal  et al., 2000). The testing material is installed between two chambers, and a fan is
installed in each chamber to mix the air (Figures). During the test, the VOC compound under investigation
is introduced into one chamber, while the initial concentration of the other chamber is zero. Partition and
diffusion coefficients are estimated based on a comparison of the measured gas-phase concentrations in
the two chambers.






X
#
<\
\
\!
1
i
1
i

X

/A\
(A

\y

///,
^^

§S

Specimen^
^^V^V

•^Membrane '

o


, x
                          Teflon Gasket
                                             V|,d \    Low concentration chamber

                                                             specimen
                                              Vj,Cj
    ^^y^  Sampling Port

High concentration chamber
Figure 8.   Schematic plot of the diffusionmetric method (Bodalal et al., 2000)

Twin dynamic chamber method
The twin chamber method features two chambers separated by the test material. One chamber is dosed
with VOC through inlet air at a constant rate, while clean air passes through the other chamber. VOC
concentrations in both chambers  are monitored continuously. Depending on the type of chamber used,
this method has several variations (Meininghaus et al., 2000,  2002; Xiong et al., 2009; Xu et al., 2012).
Figure  9 shows the generic test  facility for the twin  chamber method (Xiong el at., 2009). Different
methods are used to estimate the partition and diffusion coefficients from the experimental results. Non-
linear regression based on solutions to Fick's law is commonly used. The method proposed by Xiong et al.
(2009)  takes  into consideration the convective mass transfer although the calculation is  somewhat
complex.
                                         Page 18 of 41

-------
                      Cljn
                      material
                       Cl.in
                                                    chamber 2
                                             C2.
                                             Ci.
                                            Ci
                                                    chamber 1
                C2.mil
                                                                   Cl.o
Figure 9    Schematic plot of the dual-chamber method (Xiong et al., 2009)

Dual chamber in series method
The dual chamber in series method is a recently developed approach to estimate partition and diffusion
coefficients of SVOCs after solving issues such as low concentrations in air, difficulty of measuring the
mass change, and strong sorption effects (Liu et al., 2014). The experiment setup is presented in Figure
10, in which two environmental  chambers are operated  in series as the source and the material test
chambers. Outlet air from both chambers are measured by the polyurethane foam  (PDF) samplers. Test
materials are pre-cleaned, punched  into  circular disks, and  are mounted on  aluminum pin mounts
("buttons"), which are then placed on aluminum pin-mounted support blocks.  Each chamber contains a
cooling fan to ensure the air is well-mixed. Prior to the experiment, the test chamber walls are pre-coated
with the SVOC to be investigated. During the tests, the  material buttons are removed from the test
chamber at different exposure times to determine the amount of SVOC absorbed  by the buttons over
time. Both partition and diffusion coefficients are estimated with a degree of sorption saturation (DSS)
model, which was originally developed  by Deng et al. (2010), as the sorption saturation degree (SSD)
model.
                          Source Chamber
Test Chamber
                                 •*•— Fan

                              Sliced Caulk
           Fan

Sink Materials
                                            PUF
               PUF
Figure 10.  Schematic plot of the dual chamber method (Liu et al., 2014)
                                        Page 19 of 41

-------
Variable volume loading
The variable volume loading method uses a closed stainless steel chamber or a sealable jar. The test
specimen with known surface area and volume is placed in the chamber. Once the equilibrium condition
is reached, gas-phase concentration in the chamber is determined. The same experimental procedure is
repeated several times by changing the volume of the test specimen so the loading factor is different from
test to test. The initial concentration of the chemical in the test specimen and material-air partition
coefficient is estimated by plotting the equilibrium concentration versus the ratio of the air volume over
the volume of the test specimen.

Cup method
This method determines the solid-phase diffusion coefficient
only. Based on an ISO standard on water vapor diffusion (ISO
12572), the cup method involves a cup of  liquid VOC at
saturation in headspace. The top of the cup is covered by a
test specimen (Figure 11, Kirchner et al.,  1999; Blondeau et
al., 2003).  The system is  placed  in a  temperature  and
humidity-controlled  environment,   and   the  diffusion
coefficient of the tested specimen is estimated by weighing
the diffusion loss of VOC using a microbalance.
              Bulk ;iir phase
                 C,*=0
    Material
    sample
Liquid VOC
Porosity-based method
Diffusion coefficients can be estimated by the porosity-based
method  through  mercury  intrusion  porosimetry  tests
(Blondeau et al.,  2003). The first step is to conduct mercury
intrusion porosimetry (MIP) tests to characterize the porous
structure of the  materials of interest, followed by applying
1
'•"
- -
' f.
'?2
•:,
. ,
'•'• '
--^
.-"
C,'=CJ(T)

                              Boundary
                               layer
                                                                                   Micro balance
Figure 11.  Schematic plot of the cup
           method (Blondeau et al.,
           2003)
Carniglia's mathematical model to estimate the effective diffusivities of any gaseous species in these
materials. Porosity-based method can be applied to uniform, isotropic materials (properties are the same
in all directions within the material). However, it does not address situations where diffusion is controlled
by surface migration, which is not the case in practical building applications.

Reporting Results and Records  Retention:
A final  report shall be prepared,  and records shall be  retained  in accordance  with  40 CFR  792,
Subpart J - Records and Reports. Sampling parameters vary based on the chemical, product, and exposure
scenario of interest. All sampling parameters need to be thoroughly documented and reported:

    Initial concentration of the chemical of interest in the test material and the chamber air.
    Dimensions of the test equipment (e.g., chamber,  cup).
    Surface area, thickness, and location of the product exposed within the chamber.
    Environmental conditions: chamber air flow rates, temperatures, relative humidity values, and air
    exchange values
    Sampling and analytical methods: description or citation, including deviation from standard
    procedure. Any additional modifications to the chamber system (fans, removable sample devices,
    etc.).
    Test results: concentration vs. time  and sampling timeframe. ASTM 5116 describes a method to
    convert chamber concentrations to emission rate (in mass/time) and emission factor (in
    mass/area/time).
                                         Page 20 of 41

-------
    QA/QC data: accuracy and precision of measurements, calibrations, daily calibration checks,
    background samples, blank samples.

References:
ASTM D5116-10, 2010. Standard guide for small-scale environmental chamber determinations of
    organic emissions from indoor materials/products. ASTM International, West Conshohocken, PA.
Blondeau, P., A. Tiffonnet, A. Damian, O. Amiri and J. Molina (2003). "Assessment of Contaminant
    Diffusivities in Building Materials from Porosimetry Tests." Indoor Air 13(3): 310-318.
Cox, S. S., D. Zhao and J. C. Little (2001). "Measuring Partition and Diffusion Coefficients for Volatile
    Organic Compounds in Vinyl Flooring." Atmospheric Environment 35(22): 3823-3830.
Corsi, R., Grain, N., Fardal, J., Little, J., and Xu, Y. (2007) Determination of Sorption Parameters for 36
    VOC/Material Combinations. Final Report. EPA Report EPA 600/R-07/035-R1, submitted to Dr. Zhishi
    Guo.
Deng, Q., X. Yang and J. S. Zhang (2010). "New Indices to Evaluate Volatile Organic Compound Sorption
    Capacity of Building Materials (Rp-1321)." HVAC&R Research 16(1): 95-105.
Global CEM Net. Global Net on "Consumer Exposure Modeling". In: Kephalopoulos S., Arvanitis A., and
    Jayjock M. (Eds.). Workshop no. 2 report on "Source characterization,  transport and fate", 20-21
    June 2005, Intra (Italy), Office for Official Publication of the European Communities, Luxembourg.
    EUR 22521 EN/2, 2006b.
He, Z., W. Wei and Y. Zhang (2010). "Dynamic—Static Chamber Method for Simultaneous Measurement
    of the Diffusion and Partition Coefficients of VOCs in Barrier Layers of  Building Materials." Indoor
    and Built Environment 19(4): 465-475.
ISO (2001). "12572: 2001." Hygrothermal performance of building materials and products.
    Determination of water vapor transmission properties," AFNOR.
Kirchner, S., J. Badey, H. N. Knudsen, R. Meininghaus, D. Quenard, H. Sallee and A. Saarinen (1999).
    Sorption Capacities and Diffusion Coefficients of Indoor Surface  Materials Exposed to Vocs. Proposal
    of New Test Procedures. Indoor Air 99.
Little, J. C. and A. T. Hodgson (1996). "Strategy for Characterizing Homogeneous, Diffusion-Controlled,
    Indoor Sources and Sinks." ASTM Special Technical Publication 1287: 294-304.
Liu, X., Z. Guo and  N. F. Roache (2014a). "Experimental Method Development for Estimating Solid-Phase
    Diffusion Coefficients and Material/Air Partition Coefficients of SVOCs." Atmospheric Environment
    89: 76-84.
Liu, X., N. F. Roache and M. R. Allen (2014b). "DEVELOPMENT OF A SMALL CHAMBER METHOD FOR
    SVOC SINK EFFECT STUDY " 13th  International Conference on Indoor Air Quality and Climate, Indoor
    Air 2014; Hong Kong; Hong Kong.
Meininghaus, R., L. Gunnarsen and H. N. Knudsen (2000). "Diffusion  and Sorption of Volatile Organic
    Compounds in Building Materials-Impact on Indoor Air Quality." Environmental Science &
    Technology 34(15): 3101-3108.
Meininghaus, R. and E. Uhde (2002).  "Diffusion Studies of VOC Mixtures in a Building Material." Indoor
    Air 12(4): 215-222.
Wang, X., Y. Zhang and J. Xiong (2008). "Correlation between the Solid/Air Partition Coefficient and
    Liquid Molar Volume for Vocs in  Building Materials." Atmospheric Environment 42(33): 7768-7774.
Xiong, J., Y. Zhang, W. Yan and Z. He (2009). "An Improvement for Dynamic Twin Chamber Method to
    Measure VOC Diffusion Coefficient and Partition Coefficient." ASHRAE Transactions 115(2).
                                        Page 21 of 41

-------
Xiong, J., W. Van and Y. Zhang (2011). "Variable Volume Loading Method: A Convenient and Rapid
    Method for Measuring the Initial Emittable Concentration and Partition Coefficient of Formaldehyde
    and Other Aldehydes in Building Materials." Environmental science & technology 45(23): 10111-
    10116.
Xu, J., J. S. Zhang, X. Liu and Z. Gao (2012). "Determination of Partition and Diffusion Coefficients of
    Formaldehyde in Selected Building Materials and Impact of Relative Humidity." Journal of the Air &
    Waste Management Association 62(6): 671-679.
Zhao, D., J. C.  Little and S. S. Cox (2004). "Characterizing Polyurethane Foam as a Sink for or Source of
    Volatile Organic Compounds in Indoor Air." Journal of environmental  engineering 130(9): 983-989.
                                         Page 22 of 41

-------
Exposure Testing Protocol 4:  Participate Matter Formation  Due to
Mechanical Forces Applied to Product or Article Surfaces

Purpose:
To determine how much particulate matter is formed due to mechanical forces (abrasion) applied to the
surface of a product or article under simulated conditions designed to mimic routine use over the lifecycle
of a product.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario of interest.

Description:

Approach
Particulate matter (PM), suspended or settled,  plays an important role in human exposure to chemicals
in the indoor environment. There are three major mechanisms by which chemicals in products/articles
may transfer to particles: particle/air partitioning (sorption of vapor), particle/solid material partitioning
(migration by direct contact), and particle formation due to weathering of the source or mechanical forces
such as abrasion  applied to the source (e.g., flaking and chalking). This document describes a generic
protocol for testing particle formation due to abrasion. The design concept  of this method is based on BS
EN ISO 9073-10 (2004) and Morgeneyer et al. (2015).

Test Facility and Apparatus
The test facility, as shown in Figure 12, consists of the abrasion apparatus, test chamber (or room), particle
counters, and particle mass sampler. It is recommended that, if possible, the motor unit of the abrasion
apparatus be located outside the test chamber (Morgeneyer et al., 2015). Otherwise, particle  emission
from the motor must be checked and treated as background emissions of the test chamber.
                                                      Particle
                                                      Counters
                    Filter
         Inlet air
                     Abrasion
                    Moto Unit
                                    Abraser
                                                                       Filter
                                  Test Chamber
                                                                             Exhaust air
, Filter
                                                          Particle Mass
                                                            Sampler
Figure 12.  Schematic of the test facility for particle generation due to abrasion
                                        Page 23 of 41

-------
Abrasion Apparatus
Many standard abrasion test methods are available. In this generic protocol the Taber abrasion method
(BS EN ISO 9073-10, 2004) is recommended because it can be applied to a wide range of products/articles.
There are over 100 standard methods for Taber  abrasion tests alone.  Selection of a  proper method
depends on the type of material to be tested although the basic principles are the same.

Test Chamber
The test chamber (or room) is an air-tight enclosure with air flow, temperature and humidity controls. It
is used to house the abrasion apparatus. Typical operating conditions of the chamber are 0.3 to 0.5  air
change per hour, 23 °C, 50% relative humidity, and approximately 0.1 m/s air speed. Although several
types of enclosures can serve as the test chamber for testing particle generation, large stainless steel
chambers (ASTM D6670) are preferred as they can  meet all the aforementioned requirements.

Particle Counters
Particle counters are used to determine the size  distribution of airborne particles.  In  addition, in the
absence of valid filter samples, the results can be used to estimate the particle emission rate (described
below). In this protocol, it is recommended that the size bins of the particle counter cover the range of
aerodynamic diameters from 0.3 to 25 u.m. If a single particle counter cannot cover this range, two particle
counters with different size  ranges can be used.

Particle Mass Sampler
Particle mass monitoring allows determination of particle concentration in air and the emission rate from
the source. This is done by  collecting airborne particles  onto filters. The  particle mass is determined  by
weighing the filter before and after sampling. This method can be used to collect total suspended particles
(TSP), PM2.5 (fine particles with diameters of 2.5 u.m or smaller), PMio (particles with diameters of 10 u.m
or smaller), or inhalable coarse particles. Collecting PMio or PM2.s mass requires placing a sizing device,
most  commonly a cyclone,  upstream of the filter. For this test protocol, characterizing particle size is
recommended.

Particle mass sampling devices are commercially available. They consist of a filter sample holder, air flow
control, air pump, and timer. PTFE-coated membrane filters and quartz-fiber filters are most commonly
used for collecting particle mass from air. Because these filters are sensitive to humidity, the filters must
be conditioned under the weighing conditions before being weighed (EPA Method 201A). Dual particle
samplers should be used to  collect filter samples.

Other Equipment and Devices
A micro balance with a readability of 1 u.g or better  is needed for weighing the particle filters. The balance
should be located in a conditioned room with constant temperature and relative humidity.

Generic Test Procedure
    Before a test, prepare the test materials according to the specifications in the abrasion test.
    Condition the particle filters.
    Calibrate the flow rates  of the particle mass sampler and particle counters.
    Start the test chamber and allow the temperature, humidity, and air flow to stabilize. In the
    meantime, weight the particle filters.
    After the test chamber approaches a steady state, turn on the particle counters.
    Place the test specimen on the abrasion apparatus; start the abrasion apparatus.
    Mount a particle filter onto the particle holder.
                                         Page 24 of 41

-------
    Start the pump of the particle mass sampler.

The particle mass collected on the filter depends on the airborne particle concentration, sampling flow,
and sampling duration. For gravimetric measurements, a particle mass of 50 u.g or more is recommended;
for chemical speciation (discussed below), more mass is often needed.

Estimating the Average Particle Generation Rate from Filter Samples
The average particle generation rate can be roughly estimated from Equation 1:

R, = a=                                                                                (i)

Where Rf = Particle generation rate during abrasion test based on filter mass (u.g/h)
       Q = Air change flow rate of the test chamber (m3/h)
       m = Particle mass collected on filter (u.g)
       q  = Sampling air flow for particle mass sampler (m3/h)
       t  = Sampling duration (h)

Note that  Equation 1 underestimates the particle generation emission rate because of two factors. First,
it ignores  the particle deposition  on the interior surfaces of  the test chamber. Second, the particle
concentration in the chamber air is fairly low in the early hours because it takes time for the concentration
to reach a steady state.  The result from Equation 1 can  be corrected for these factors by means of
mathematical modeling if the air change rate of the chamber and particle deposition rate, which is size
dependent, are known.

Estimating the Average Particle Generation Rate Based on Data from Particle Counters
If the filter sampler cannot collect enough particle mass, the  particle generation rate can be roughly
estimated based  on the data from particle counters (Equation 2):

^  _ 	Q K P V?   ffif. _  JVri-)  d?                                                   (2)
  c        6q       i-ov  i      ou   i

Where Rc = Particle generation rate during abrasion test based on data from particle counter (u.g/h)
       Q  = Air change flow rate of the test chamber (m3/h)
       p  = Particle density (g/cm3)
       q  = Sampling air flow for particle counter (m3/h)
       NI = Particle number count in the ith size bin during abrasion test
       A/o/= Particle number count in the ith size bin for chamber background
       d,  = Geometric mean diameter for the ith size bin (u.m)
       n  = Number of size bins.

Like Equation 1,  Equation 2  is also subject to correction for particle deposition and non-steady-state
condition in early hours.

Chemical  Speciation
The filter  samples can be further analyzed for the  chemical  composition of the particles, known as
speciation. A wide range of physical and chemical methods are  available for particle speciation. Method
selection depends on the chemical, article, and exposure scenario of interest. It is beyond the scope of
this protocol to discuss technical details about particle speciation.
                                         Page 25 of 41

-------
Safety Issue
It is highly recommended the abrasion apparatus be operated remotely outside the test chamber. If the
operator must be inside the chamber during the test, a safety and health plan must be developed and
implemented.

Reporting Results and Records Retention:
A final  report shall be  prepared, and records  shall  be retained in accordance  with  40 CFR  792,
Subpart J - Records and Reports. For example, the following key information should be included in the
report:

    Test material: material name, intended use, uniformity (homogeneous, layered, spray application,
    coating, etc.), and dimensions of test specimens.
    Abrasion apparatus: abrader brand and model number, abrading type (abrasive characteristics of
    the wheel), and operating parameters.
    Test chamber: chamber brand and model number, volume, dimensions, and interior surface
    material.
    Environmental conditions: chamber air flow rate, temperature, relative humidity, and air speed
    expressed in  arithmetic mean and standard deviation.
    Particle counters: particle counter type, brand, and model number.
    Particle mass sampler: sampler brand and model number, filter type and size, sampling flow rate
    and duration.
    Test procedure: description or citation, including deviation from standard procedure.
    Test results: particle counts vs time for each size bin and sampling air flow; gravimetric data for
    particle mass, sampling airflow and sampling duration.
    QA/QC data:  accuracy and precision of measurements, calibrations, daily calibration checks,
    background samples, blank samples.

References:
ASTM 6670-01 (2007) Standard  Practice for Full-Scale Chamber Determination of Volatile Organic
    Emissions from  Indoor Materials/Products, available at http://www.astm.org/Standards/D6670.htm
BS EN ISO 9073-10 (2004). Lint and other particles generation in the dry state., available at
    http://shop.bsigroup.com/ProductDetail/?pid=000000000030099719
Morgeneyer, M.,  Shandilya, N., Chen, Y.M., Bihan, O.L (2015). Use of a modified Taber abrasion
    apparatus for investigating the complete stress state during abrasion and in-process wear particle
    aerosol generation, Chemical Engineering Research and Design, 93:251-256.
                                        Page 26 of 41

-------
Exposure Testing Protocol 5:  Photolysis under Simulated Indoor
Lighting Conditions

Purpose:
To determine whether a chemical in a product or article is subject to photolytic degradation under
simulated indoor lighting conditions and what the major degradation products are.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario of interest.
Description:

(1)  Approach
Photolysis, or photolytic degradation, is a chemical reaction by which the compound is broken down by
light (photons). This process is relevant to indoor environmental quality and human exposure because, in
some  cases, the broken-down chemicals may be hazardous. While most chemicals found  in indoor
products/articles are  expected to be resistant to photolysis  under indoor lighting  conditions, a few
chemicals are not.  For example, decabrominated diphenyl ethers, or decaBDE, is known to undergo
photolytic  debromination  under natural  sunlight, forming less brominated congeners (Stapleton and
Dodder, 2008).  It is less clear, however, how significant decaBDE photolysis is under indoor lighting
conditions.

In this protocol, a generic method is described for  testing the photolysis potential  for chemicals like
decaBDE by exposing the  test material to simulated sunlight through windows in an accelerated
weathering chamber and potential photolysis products are detected from air (by static air sampling), the
surface of the test specimens (by wipe sampling), and settled dust (by dust sampling). The presence or
absence of photolysis products  in the samples can be determined qualitatively  by comparing the
chromatograms and  quantitative analytical results for exposed samples with those for unexposed
samples.

(2)  Facility and Apparatus

Test Chamber
Photolysis tests should be conducted in an accelerated weathering chamber, which provides ultraviolet
(UV) irradiation, controlled temperature, and  humidity. Two types of weathering  chambers are
commercially available (ASTM  G154 and ASTM G155).  Those that conform  to  ASTM G155 are
recommended for this protocol. To simulate indoor lighting conditions, the system must have optical
filters that generates sunlight through window glass  (ASTM D 4459-06: Standard Practice for Xenon-Arc
Exposure of Plastics Intended for Indoor Applications). A chamber system conforming to ASTM D 4459-06
can provide spectral  irradiance of approximately 0.3 (W/m2/nm) at 340 nm when operated  in the
continuous  light-on mode without water spray. This light source satisfies the light intensity requirement
of 5 W/m2 over the test specimens. For testing settled dust, the chamber model must allow the panels to
be placed on a horizontal (or nearly horizontal) tray.
                                        Page 27 of 41

-------
Note that the standard methods for accelerated  aging tests under UV  irradiation are intended for
examining the changes of physical properties. To detect photolysis products, the test procedure requires
several modifications and additional steps, as described below.

Passive Air Sampler
Passive air samplers are used to capture chemical  vapors emitted from the test specimens during the
accelerated  weathering  test. This  method  determined  time-averaged  concentrations  by  using
polyurethane foam disk as the sampling  media (Harrad & Abdallah, 2008). The sampler can be mounted
onto the chamber walls prior to a test. The sample media removed from the chamber can be extracted by
solvents and analyzed  for  potential  photolysis  products. The  analytical  procedure depends on the
properties of the target chemicals.

(3)  Test Specimens
The product or article to be tested are cut into panels. Different chambers may have different standard
panel sizes and some chambers allow custom-size panels. In general, panels for wipe sampling should be
at least 100 cm2 in size and those for testing  settled dust at least 500 cm2. For a given product or article,
12 panels are needed for wipe sampling and 12 for testing settled dust.

(4)  Wipe Sampling
If photolysis products are present on the exposed surface of the test specimens, they can be collected by
wipe sampling.

Wipe Sampling on Solid Surfaces
ASTM D 6661-10, Standard Practice for Field Collection of Organic Compounds from Surfaces Using Wipe
Sampling, or an equivalent method, shall be used for surface sampling on solid panels. The wipe samples
shall be extracted and then analyzed for potential photolysis  products.

Surface sampling on Fabric Swatches
The method  is  based on the California roller method  (Ross et  al., 1991;  Fuller  et  al., 2001)  with
modifications. Use 3" x 6" heavy filter paper instead of cotton gauze pad;  place the fabric swatch on a
pre-cleaned, non-porous, flat surface (such as a rigid metal plate or polished granite block); place the
heptane-wetted filter paper on the fabric swatch; place a 3" x 6" stainless  steel (or aluminum) plate on
the paper filter; add additional weights on the plate such that the total weight is 2 pounds (Ib); wait for 5
minutes; remove plate and weights; extract the paper filter.

(5)  Dust Sampling
Photolysis may be difficult to detect on product or article surfaces by wipe sampling, and tests with settled
dust are recommended.

House Dust or Surrogate Dust
Ideally, cleaned-up standard house dust should be used (Stapleton and Dodder, 2008). Because of high
cost of the standard reference material,  surrogate  dust (e.g.,  Arizona test dust) can be used. In either
cases, the dust must be free of the chemical of interest and its degradation products. Otherwise, the dust
must be cleaned by solvent extraction. For Arizona test dust,  10-u.m mean diameter is recommended.

Dust Application
Larger test specimens (i.e., coupons), such as  6" x 6" (15.2  cm x  15.2 cm), are recommended for tests with
settled dust. The goal is to apply an adequate amount of  house dust (or surrogate dust) on the coupons
                                         Page 28 of 41

-------
without forming a thick layer of dust. The target dust load is between 3.0 to 4.3 mg/cm2 coupon, which is
roughly equivalent to 0.7 to 0.9 g dust per panel.

The house dust can be deposited on test  specimens by using a separate dust  deposition chamber
(O'Shaughnessy et al., 2002) or spiked manually on test specimens (Ashley et al., 2007).

Dust Sampling
Dust samples over the test coupons will be collected by the micro-vacuuming method (ASTM D 7144-05a;
Ashley etal., 2007).

(6)  Analytical Methods
Selection of the analytical methods for air,  wipe, and dust samples depends on the properties of the
chemicals of  interest  and the type of sampling media. For example, for photolysis of decaBDE,
chromatography or mass spectrometry in electron capture negative ionization mode (GC/MS-ECNI) has
been used (Stapleton et al., 2008). Identification and quantification of photolysis products in the samples
are sometimes challenging because  of the dominance of the chemical of interest (i.e.,  the  parent
compound) in the chromatograms. This issue can  be resolved by using highly sensitive instrument and by
adopting a pre-separation method such as preparative chromatography.
(7)  Generic Procedure for Photolysis without Dust
    Prepare 12 3" x 6" (7.6 cm x 15.2 cm) coupons (panels
    or fabric swatches).
    Take wipe samples on 3 coupons, which represent no-
    exposure conditions.
    Clean the interior surfaces, wherever reachable, and
    the sample tray of the test chamber by washing with
    soap and water, wiping with toluene, and wiping with
    methanol.
    Take two wipe  samples (100  cm2 each) from the
    chamber walls.
    Place three  passive air samplers  (PDF disks) on the
    supporting cradle about half chamber height.
    Place the remaining 9 coupons on the sample tray.
    Close the chamber door, set the temperature at 35 °C
    and  relative humidity at 30% (The moisture content is
    roughly equivalent to that of 50% RH at 25 °C).
    Turn on the  UV light to start the test.
    On day 4, turn off the UV light,  open the chamber
    door, and perform the following steps:
       Remove three coupons from the chamber for
       taking wipe samples.
       Remove one PDF disk for determination of time-
       integrated  air  concentrations of  the  target
       chemical and potential photolysis products.
       Close the chamber door and turn on the UV light
       to restart the test.
    On day 15 and day 30, repeat the steps on day 4.
   control    day 4    day IS   day 30
             I
                              PUF disks
              	
                       •—p»temp: 35
                          L-»RH: 30%
                          L«UV light
Figure 13. Graphic example of generic
          procedure for photolysis
          without dust
                                        Page 29 of 41

-------
After all samples are removed, take two wipe samples from the chamber walls (100 cm2 each).
                                                              day 4
• dust sample (x3)

day IS    day 30
                                                                               PUF disks
(8)   Generic Procedure for Photolysis Test with Dust
-   Prepare  12  6"  x 6" (15.2 cm  x 15.2  cm) coupons
    (panels or swatches).
    Collect triplicate dust samples  for determination of
    background  concentrations.
    Clean  the interior surfaces, wherever  is reach-able,
    and the sample tray of the test chamber by washing
    with soap and water, wiping with toluene, and wiping
    with methanol.
    Take  two wipe samples (100 m2 each)  from the
    chamber walls.
    Place three  passive air samplers (PUF  disks)  on the
    supporting cradle about half chamber height.
    Apply  test  dust onto  coupons according  to the
    method described in Section (5).
    Take wipe samples on 3 coupons, which represent no-
    exposure conditions.
    Place the remaining 9 dust-loaded coupons  on the
    sample tray.
    Close the chamber door, set the temperature at 55 °C.
    Do not use water spray for humidity control because
    water   droplets  falling  onto  dust-loaded  test
    specimens may complicate the  interpretation of test
    results.
    Turn on the  UV  light to start the test.
    On day 4, turn  off the UV light,  open the chamber
    door, and perform the following steps:
       Remove three coupons from the chamber  for
       collecting dust samples according to Section (5).
       Remove one PUF disk for determination of time-
       integrated  air  concentrations  of the  target
       chemical and potential photolysis products.
       Close the chamber door and turn on the UV light to restart the test.
    On day 15 and day 30, repeat the steps on day 4.
    After all samples are removed, take two wipe samples from chamber walls (100 cm2 each).

(9)   General Procedure for Tests without UV-light
If the test results from steps (7) and (8) suggest the presence of photolysis products in air, surface or dust
samples, it is recommended to conduct a test without the UV-light (i.e., dark cycle). This is done by
following steps (7) and (8) except that the UV light is turned off.

Reporting Results and Records Retention:
A final report shall  be  prepared,  and  records  shall  be  retained in accordance with 40 CFR 792,
Subpart J - Records and Reports.
                                                                            >temp:S5°C
                                                                           -•RH: NONE
                                                                           -*UV light
                                                 Figure 14. Graphic example of generic
                                                           procedure for photolysis with
                                                           dust
                                     Page 30 of 41

-------
The standard test methods mentioned above contain sections for reporting. For example, key
information to be reported includes:

    Test material: material name, intended use, uniformity (homogeneous, layered, spray application,
    coating, etc.), dimensions of test specimens, exposed area, treatment of sample edges (sealed or
    exposed) and information about sample creation, transport, and storage.
    Target chemical(s) and their basic properties: CAS number, molecular formula, vapor pressure,
    chemical reactivity, concentration in material, etc.
    Test chamber: chamber type, model name, volume, dimensions, and interior surface material.
    Test procedure: description or citation, including deviation from standard procedure.
    Sampling methods for air, wipe, and dust samples and analytical methods — description or citation,
    including deviation from standard procedure. Description of accuracy and precision
    Analytical methods: description or citation, including deviation from standard procedure.
    Environmental conditions: lighting conditions (lamps, optical filter, light spectrum and intensity),
    chamber temperature (expressed in arithmetic mean and standard deviation), and moisture content
    in cooling air.
    Test results: chromatograms of air, wipe and dust samples, identification of peaks, time-averaged
    concentrations in chamber air from static air sampler, concentrations in wipe and dust samples.
    QA/QC data: accuracy and precision of measurements, calibrations, daily calibration checks,
    background samples, blank samples.

References:
Ashley, K., Applegate, G.T., Wise, T.J., Fernback, J.E., and Goldcamp, M.J. (2007). Evaluation of a
    standardized micro-vacuum sampling method for collection of surface  dust. Journal of Occupational
    and Environnemental Hygiene. 4:215-223.
ASTM D7144 - 05a (2011) Standard Practice for Collection of Surface Dust by Micro-vacuum Sampling for
    Subsequent Metals Determination
ASTM G154-12a, Standard  Practice for Operating Fluorescent Ultraviolet (UV) Lamp Apparatus for
    Exposure of Nonmetallic Materials, ASTM International, West Conshohocken, PA, 2012,
    www.astm.org
ASTM G155-13, Standard Practice for Operating Xenon Arc Light Apparatus for Exposure of Non-Metallic
    Materials, ASTM International, West Conshohocken, PA, 2013, www.astm.org
Harrad, S. and Abdallah, M.A. (2008). Calibration of two passive air sampler configurations for
    monitoring concentrations of hexabromocyclododecanes in indoor air. Journal of Environmental
    Monitoring, 10(4):527-31.
O'Shaughnessy, P., Svendsen, E., Thorne, P.S., and Reynolds, S. (2002). A dust-settling chamber for
    sampling-instrument comparison studies. Presented at American Industrial Hygiene Association
    Conference and Exposition, San Diego, CA, June 1-6, 2002.
Ross, J., Fong, H., Thongsinthusak, T., Margetich, S., and Krieger,  R. (1991).  Measuring potential dermal
    transfer of surface pesticide residue generated from indoor fogger use: using the CDFA roller
    method interim report II. Chemosphere. 22:975-984.
Stapleton, H.M., Kelly, S.M,; Allen, J.G., McClean, M.D., and Webster, T.F. (2008). Measurement of
    polybrominated diphenyl ethers on hand wipes: estimating exposure from hand-to-mouth contact.
    Environmental Science & Technology. 42:3329-3334. 2008.
Stapleton, H.M. and Dodder, N.G. (2008). Photodegradation of decabromodiphenyl ether in house dust
    by natural sunlight. Environmental Toxicology and Chemistry. 27:306-312.
                                         Page 31 of 41

-------
Exposure Testing Protocol 6: Oral  Exposure - Migration Rate and
Transfer Efficiency

Purpose:
To collect information on how much of a chemical migrates from an article or material into simulated
saliva overtime.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario  of interest.

Description:

Approach
The methods for measuring migration from  articles into simulated  saliva have been  described by the
European Commission Joint Research Centre (JRC) (Simoneau et al., 2001), and several other studies have
also used this approach to estimate migration rates of chemicals into saliva (Bouma and Schakel 2001)
(Bouma et al 2002) (Corea-Tellez et al 2008) (Earls  et al 2003) (Masuck et al 2011) (Niino et al 2001) (Niino
et al 2002)  (Ninno et al 2003) (Ozer and Gucer 2011) (Simoneau  et al 2009) (TNO  Nutrition and Food
Research 2001). The U.S. Consumer Product Safety Commission (CPSC) recently characterized exposure
of phthalates, including mouthing, using measured migration rates (Babich 2014). The head over heels
(HOH) approach, also referred to as aggressive agitation, measures the amount of chemical that migrates
from an article into  simulated saliva. This migration is typically reported in u.g/10 cm2/hour. Migration
rates quantify the  rate at which a chemical that is  a part of the article itself migrates from an article over
time. Additional information that characterizes  the duration of the experiment and expected conditions
of use, such as duration of mouthing time for the article or material is also needed to estimate exposure.

The transfer of a chemical deposited on the surface of an article onto hands and the transfer of a chemical
from the surface  of an article to the mouth are defined as the  hand-to-mouth and object-to-mouth
transfer efficiencies, respectively. The transfer efficiency likely varies  based on the type of material, level
of surface loading, and the physical form  of the chemical  itself (liquid or solid). A recently published
transfer efficiency database contains all publicly available transfer efficiency values and includes a
discussion of methods for measuring oral saliva transfer efficiency in  the Gorman NG et al. (2012) paper.
Both the  hand-to-mouth and object-to-mouth transfer efficiencies are important in characterizing dust
ingestion. Additional information that characterizes the frequency of hand-to-mouth and object-to-mouth
transfers  is needed for the article or material of interest in order to estimate intake.

Preparation of Saliva
There are various  approaches to prepare artificial saliva.  It is recommended that saliva is prepared at a
representative temperature and pH and contain relevant enzymes and salts in concentrations likely to be
present within the human mouth. The composition of the saliva as well as the testing conditions of the
saliva within the Head over Heels (HOH) testing apparatus should be transparent and well documented.
An in vitro  model was developed to estimate extraction via saliva (Brandon et al 2006). That paper
references a composition of saliva from Versantvoort et al (2005) which is presented below. Another
recent paper (Marques et al 2011), provides five different  approaches to simulate saliva.
                                        Page 32 of 41

-------
Versantvoort et al 2005
    Inorganic Solution:  10 ml of 89.6 g/L KCI solution,
                      10 ml of 20 g/L KSCN solution,
                      10 ml of 88.8 g/L NaH2PO4 solution,
                      1.7 ml of 175.3 g/L NaCI solution, and
                      20mLof84.7g/LNaHCO3
    Organic Solution:   8 ml of 25 g/L urea solution
    Add to Inorganic and Organic Solution:
290 mg alpha-amylase,
15 mg uric acid, and
25 mg mucin
-   pH6.8+/-0.2

Marques et al 2011
-   Simulated Saliva 1:  0.72 g/L KG,
                      0.22 g/L calcium chloride dihydrate,
                      0.6 g/L NaCI,
                      0.68 g/L potassium phosphate monobasic,
                      0.866 g/L sodium phosphate dibasic (dodecahydrate),
                      1.5 g/L potassium bicarbonate,
                      0.06 g/L potassium thiocyanate, and
                      0.03 g/L citric acid
                      (pH6.5)
-   Simulated Saliva 2:  0.72 g/L KG,
                      0.22 g/L calcium chloride dihydrate,
                      0.6 g/L NaCI,
                      0.68 g/L potassium phosphate monobasic,
                      0.866 g/L sodium phosphate dibasic (dodecahydrate),
                      1.5 g/L potassium bicarbonate,
                      0.06 g/L potassium thiocyanate, and
                      0.03 g/L citric acid
                      (pH 7.4)
    Simulated Saliva 3:  0.228 g/L calcium chloride dihydrate,
                      1.017 g/L NaCI,
                      0.204 g/L sodium phosphate dibasic (heptahydrate),
                      0.061 g/L magnesium chloride hexahydrate,
                      0.603 g/L potassium carbonate hemihydrate,
                      0.273 g/L sodium phosphate monobasic monohydrate,
                      1 g/L submaxillary mucin, and
                      2 g/L alpha-amylase
-   Simulated Saliva 4:  0.149 g/L KG,
                      0.117 g/L NaCI,
                      2.1 g/L sodium bicarbonate,
                      2 g/L alpha-amylase, and
                      1 g/L mucin gastric
-   Simulated Saliva 5:  8.0 g/L NaCI,
                      0.19 g/L potassium phosphate monobasic, and
                                         Page 33 of 41

-------
                      2.38 g/L sodium phosphate dibasic
                      (pH6.8)

Preparation of Samples, Extraction, and Analysis
To prepare samples, discs, coupons, or circular samples are cut from the surface of the test article. The
diameter of the samples should be approximately 2 inches.

Note, for each extraction, 50 ml is typically used. The weight and the volume of the simulated saliva
should be reported. Many test procedures and ASTM F963 require a 50:1 ratio of solvent to sample for
this type of extraction. The samples will be extracted four times each in 50 ml of simulated saliva in a 250
ml Schott Duran (or similar) bottle for 30 minutes. The bottle is shaken at 60 rpms in a circular head-over-
heels (HOH) motion for the duration of the experiment, vertical diameter of 2 feet.

The liquid simulated saliva extract is removed after each extraction and saved for analysis. Afresh 25 ml
of simulated saliva is added to the bottle containing the sample, and the bottle is shaken as above for 30
minutes. The replicate simulated saliva extract is then removed and also saved for analysis. The HOH
procedure is then repeated a third time. Each separate solution obtained from these shakings is analyzed
for the chemical of interest.
                                                                    60 rpm
                                                                    30 minutes
                                                                    x4
                                       50 ml simulated saliva
Figure 15.  Graphic example of procedure for analyzing migration from product or article surface
           to saliva

For chemical analysis, 10 ml of the simulated saliva is placed in a test tube. One ml of xylene (or suitable
solvent) is added to the test tube and the tube is spun for one minute. The supernatant solvent is analyzed
for chemical content by injecting 1.0 ul into the Gas Chromatography/Mass Spectrometry (GC/MS, or
other suitable instrument). The results for the four extractions are then combined. The chemicals present
in simulated saliva will be analyzed using different analytical methods, depending on the chemical present.

A variety of analytical methods can be used depending on the chemical(s) present. For example, GC/MS,
inductively  coupled plasma atomic emission spectroscopy (ICP) or HPLC. Note the  instrumentation
conditions for whichever analytical technique is used.
                                         Page 34 of 41

-------
Reporting Results and Records Retention:
A final  report shall be prepared,  and records shall be  retained in accordance  with  40 CFR  792,
Subpart J - Records and Reports.

For example, key information to be reported includes:

   Sampling and analytical methods — description or citation, including deviation from standard
   procedure, if applicable.
   Description of simulated saliva composition (components, weight, volume)
   Description of tested material (size, dimensions)
   QA/QC data: accuracy and precision of measurements

References:
Babich, M. (2014). Report to the U.S. Consumer Product Safety Commission by the CHRONIC HAZARD
   ADVISORY PANEL ON PHTHALATES AND PHTHALATE ALTERNATIVES. Appendix E-2-4, available at
   http://www.cpsc.gov/PageFiles/169914/Appendix-E2-Substitutes-Exposure-FINAL.pdf
Bhooshan, B., (2005). Vinylidene Chloride Testing in Mattress Barrier Samples. Tab H.  Pages 535-537,
   available at https://www.nvwa.nl/txmpub/files/?p  file  id=10485
Bouma K and Schakel D.J., 2001. Plasticisers in Soft PVCToys. Report number: NDTOY002/01, available
   at https://www.nvwa.nl/txmpub/files/?p file id=10485
Bouma, K., & Schakel, D. J. (2002). Migration of phthalatesfrom PVCtoys into saliva simulant by
   dynamic extraction. Food Additives & Contaminants, 19(6), 602-610., available at
   http://www.tandfonline.com/doi/abs/10.1080/02652030210125137tf.VgqkTJe2q8R
Brandon, E. F., Oomen, A. G., Rompelberg, C. J., Versantvoort, C. H., van Engelen, J. G., & Sips, A. J.
   (2006). Consumer product in vitro digestion model: Bioaccessibility of contaminants and its
   application in risk assessment. Regulatory Toxicology and Pharmacology, 44(2), 161-171.
Cobb, D. (2005). Migration of Flame retardant Chemicals in Mattress Barriers. Tab H. Pages 542-553,
   available at http://www.cpsc.gov/PageFiles/88231/matttabh.pdf
Corea-Tellez, K. S., Bustamante-Montes, P., Garcia-Fabila, M., Hernandez-Valero, M. A., & Vazquez-
   Moreno, F. (2008). Estimated risks of water and saliva contamination by phthalate diffusion from
   plasticized  polyvinyl chloride. Journal of environmental health, 71(3). 34-9., available at
   http://europepmc.org/abstract/med/18990931
Earls, A. O., Axford, I. P., & Braybrook, J. H.  (2003). Gas chromatography-mass spectrometry
   determination of the migration of phthalate plasticisers from polyvinyl chloride toys and childcare
   articles. Journal of chromatography A, 983(1). 237-246.. available at
   http://www.sciencedirect.com/science/article/pii/S0021967302017363
Masuck, I., Hutzler, C., & Luch, A. (2011). Estimation of dermal and oral exposure of Children to scented
   toys: Analysis of the migration of fragrance allergens by dynamic headspace GC-MS. Journal of
   separation science, 34(19), 2686-2696., available at
   http://onlinelibrarv.wiley.com/doi/10.1002/issc.201100360/abstract;isessionid=CD01BF2ED5C27FO
   3C6AClE405B78E4D6.f03t04?deniedAccessCustomisedMessage=&userlsAuthenticated=false
Marques, M. R., Loebenberg, R., & Almukainzi, M. (2011). Simulated biological fluids with possible
   application in dissolution testing. Dissolution Technol,18(3), 15-28 (Table 10).
Ng, M. G., Semple, S., Cherrie, J. W., Christopher, Y., Northage, C., Tielemans, E., Veroughstraete, V., &
   Van Tongeren, M. (2012). The relationship between inadvertent ingestion and dermal exposure
                                         Page 35 of 41

-------
    pathways: A new integrated conceptual model and a database of dermal and oral transfer
    efficiencies. Annals of occupational hygiene.
Niino T, Ishibashi T, Itho T, Sakai S, Ishiwata H, Yamada T, and Onodera S, 2001. Monoester formation
    by hydrolysis of dialkyl phthalate migrating from PVC products in human saliva. Journal of Health
    Science, 47(31:318-322.. available at http://ihs.pharm.or.jp/data/47%283%29/47%283%29p318.pdf
Niino T, Ishibashi T, Itho T, Sakai S, Ishiwata H, Yamada T, Onodera S, 2002.  Comparison of Diisonoyl
    Phthalate Migration from Polyvinyl chloride products into human saliva in Vivo and into saliva
    simulant In vitro. Journal of Health Science, 48(3):227-281., available at
    http://ihs.pharm.or.jp/data/48%283%29/48  277.pdf?origin=publication detail
Niino T, Asakura T, Ishibashi T, Itho T, Sakai S, Ishiwata H, Yamada T, Onodera S, 2003. A Simple and
    Reproducible Testing Method for Dialkyl Phthalate Migration from Polyvinyl Chloride Products into
    Saliva Simulant. J. Food Hyg. Soc. Japan Vol. 44, No. 1, available at
    https://www.istage.ist.go.ip/article/shokueishi/44/l/44  1 13/  pdf
Ozer, E. T., & Gucer, S. (2011). Determination of some phthalate acid esters in artificial saliva by gas
    chromatography-mass spectrometry after activated carbon enrichment. Talanta, 84(2). 362-367..
    available at http://www.sciencedirect.com/science/article/pii/S003991401100035X
Simoneau, C, Hannaert, P., and Sarigiannis, D. E.  2009. Effect of the nature and concentration of
    phthalates on their migration from PVC materials under dynamic simulated conditions of mouthing.
    EUR 23813 EN; pp. 20..available at
    http://publications.irc.ec.europa.eu/repository/bitstream/JRC51604/reqno jrc51604 chemtest par
    t2-phthalates release toys  cs02009 05 26.pdf%5Bl%5D.pdf
Thomas, T. A., & Brundage, P. M. (2006). Quantitative Assessment of Potential  Health Effects from the
    Use of Fire Retardant Chemicals in Mattresses. CPSC. Tab D. Section 2. Methods. Pages 10 -13,
    available at http://www.cpsc.gov/PageFiles/88208/matttabd.pdf
Thomas, T. A., & Brundage, P. M. (2006). Quantitative Assessment of Potential  Health Effects from the
    Use of Fire Retardant Chemicals in Mattresses. CPSC. Tab D. Appendix 2. Experimental Protocol Flow
    Chart. Pages 239-242, available at http://www.cpsc.gov/PageFiles/88208/matttabd.pdf
TNO Nutrition and Food Research, 2001.  Migration of phthalate plasticizers from soft PVC toys and
    childcare articles -Final report. Netherlands Organization for  Applied Scientific Research, available
    at http://ec.europa.eu/enterprise/sectors/chemicals/files/studies/phthalates2 en.pdf
Versantvoort, C. H., Oomen, A. G., Van de Kamp,  E., Rompelberg, C. J., & Sips, A. J. (2005). Applicability
    of an in vitro digestion model in assessing the bioaccessibility  of mycotoxins from food. Food and
    Chemical Toxicology,43(l), 31-40.
                                         Page 36 of 41

-------
Exposure Testing Protocol 7: Dermal Exposure -  Potential Exposure

Purpose:
To collect information on how much of a chemical load is present on the surface of the skin and potentially
available for exposure through various exposure pathways.

Modifications:
This protocol is general, and it is anticipated that during protocol development and finalization,
additional modifications will be made to tailor the sampling parameters or analytical techniques to the
specific chemical and product tested. It is anticipated that during protocol development, Agency
recommendations will be incorporated to tailor sampling parameters or analytical techniques to the
specific product, chemical, and exposure scenario of interest.
Description:
For this protocol, actual skin (i.e. animal skin, cadaver skin, human subject skin, where proper ethical and
scientific research requirements have been met, etc.) or a skin surrogate such as filter paper may be used
to estimate chemical load present on the surface of the skin. If human subjects are used for the testing,
ensure that all requirements related to issues associated with scientific and ethical aspects of human
subject research are adhered to. There are three primary mechanisms for chemical loading on  to the
surface of the skin:

   Application of liquid or semi-solid of the chemical as part of a formulation
   Contact with surface of article or building material and migration into simulated sweat.
   Proximity of skin to vapor-phase chemical concentrations in the air
The first mechanism applies primarily to products; the second to articles. The third mechanism can occur
as a result of product use or article exposure. This exposure pathway may also be significant depending
on the chemical, product, and environment of interest for the exposure scenario.  Methods, such as those
presented by Weschler et al, 2015 and Gong et al,  2014, show promise, and information from studies like
this could inform an expanded basis for protocols characterizing the dermal pathway in the future.

Note, potential dermal exposure is described here. There are approaches available to estimate absorbed
dose if dermal exposure is expected to be an important exposure pathway but this is outside the scope of
this protocol.  In vivo  measurements, in vitro  measurements and/or measured permeability coefficients
could be used to estimate absorbed dose. The flux of a chemical across the skin membrane, whether an
infinite or finite dose is assumed, exposure duration, and comparison of different approaches can be
considered (OECD 2004a) (OECD 2004b) (Buist et al 2010) (Frasch et al 2014).

Approach  for Determination of Film Thickness from Application of Liquids or Semi-solid Product
For products in which skin contact other than direct application to the skin occurs, the measurement of
the thickness of the product film that remains on  the skin after contact is used to characterize the mass
of product that remains on the  skin after contact. The film thickness can be measured in up to five use
scenarios depending  on the  intended product use. In all scenarios, the product should be prepared
according to use instructions. Because the test is measuring the thickness of a film of a product retained
on the skin as a result of use of the product, a surrogate test product with similar properties (e.g., volatility,
                                        Page 37 of 41

-------
viscosity, etc.) can be used for testing. For example, surrogate test products which are generally regarded
as non-toxic or safe should be used if human subjects will be used during testing.

Primary and secondary contact
For the initial contact scenario, a cloth saturated with the product should be rubbed over the front and
back of both clean, dry hands. For the secondary contact scenario, a cloth saturated with the product
should be rubbed over the front and back of both hands for a second time, after as much as possible of
the liquid that adhered to skin during the first contact event was removed using a clean cloth. The subject's
hands should then be fully wiped, defined as wiping with a clean dry cloth as thoroughly as possible for
10 seconds. The film thickness is determined by dividing the  difference in cloth weight before and after
wiping by the surface area of the hand and the density of the prepared product. Four to 6 replicate tests
should be conducted  and reported.

Immersion
To measure the film thickness that results from immersion, the hand should be immersed in the prepared
product and then allowed to drip back into the container for  30 seconds. The weight of the container of
prepared product should be  weighed before and after  immersion. The difference in weight of the
container divided by the surface area of the hand, normalized by the density of the product is the film
thickness. Four to 6 replicate tests should be conducted and reported.

Contact from Handling a Wet Rag
To estimate film thickness from handling a wet rag, a cloth saturated  with the product should be rubbed
over the palms of both hands in a manner simulating handling of a wet cloth. The subject's hands should
then be fully wiped, defined as wiping with a clean dry cloth as thoroughly as possible for 10 seconds. The
film thickness is  determined by dividing the difference in cloth  weight before and after wiping by the
surface area of the hand and the density of the prepared product. Four to 6 replicate tests should be
conducted and reported.

Contact from Cleaning a Spill
A subject should  use a  clean cloth to wipe  up 50  ml of prepared product poured onto a non-porous
surface. After cleanup, the subject's hands should then be fully wiped, defined as wiping with a clean dry
cloth as thoroughly as possible for 10 seconds. The film thickness  is determined by dividing the difference
in  cloth weight before and after wiping by the surface area of the hand and the density of the prepared
product. Four to 6 replicate tests should be conducted and reported.

Approach for Estimating Migration into Simulated Sweat from Contact with an Article
A small or large scale experiment can be used to evaluate sweat facilitated migration from an article
onto skin or skin surrogate.  The sampling conditions shall be varied based on the chemical, article, and
scenario of interest. Parameters that shall be varied include the:

    size and thickness of the article,
    amount of surrogate sweat applied,
    amount and timing of pressure (psi) applied,
    size of skin surrogate material used, and
    additional barrier present or not present between article surface and surrogate skin  material of filter
    paper
                                         Page 38 of 41

-------
Preparation of Sweat
Artificial sweat or perspiration is the reagent extract solution. Typical components of artificial sweat
include water, lactate, urea, sodium, potassium, calcium, and magnesium. Saline solution may be used as
a starting  point.  Because the swelling of water-soluble polymers is suppressed by some metal ions,
especially  calcium, and by low pH, Marques et al. (2011) provide five different approaches to simulate
sweat with different concentrations of calcium and different pH.

    Simulated Sweat 1 (3 milliequivalents of calcium ions): 2.92 mEq/L NaCI,
                                                     0.166mEq/LCaCI2,
                                                     0.12 mEq/L MgSO4, and
                                                     1.02 mEq/L potassium phosphate monobasic
                                                     (pH5.4)
    Simulated Sweat 2 (60 milliequivalents of calcium ions):5.49 mEq/L NaCI,
                                                     3.32mEq/LCaCI2,
                                                     0.24 mEq/L MgSO4, and
                                                     1.36 mEq/L potassium phosphate monobasic
                                                     (pH4.5)
    Simulated Sweat 3 (120 milliequivalents of calcium ions): 5.49 mEq/L NaCI,
                                                     6.64 mEq/L CaCI2,
                                                     0.24 mEq/L MgSO4, and
                                                     1.36 mEq/L potassium phosphate monobasic
                                                     (pH4.5)
    Simulated Sweat 4 (240 milliequivalents of calcium ions): 5.49 mEq/L NaCI,
                                                     13.28mEq/LCaCI2, and
                                                     0.24 mEq/L MgSO4, 1.36 mEq/L potassium
                                                     phosphate monobasic (pH 4.5)
-   Simulated Sweat 5: 0.5 % (in mass) NaCI,
                     0.1 % lactic acid, and
                     0.1 % urea with the recommended volume of simulated fluid (about 1 mL per
                     cm2 sample area)

Preparation of Samples, Extraction, and Analysis
A small-scale experiment can be used to evaluate an article coupon with surface area corresponding to a
circle with a diameter of 5.5 cm. The article sample is placed in a 600 mL beaker and covered with skin or
skin surrogate such as Whatman #2 filter paper large enough to cover the article sample. Two to 4 mL of
simulated  sweat extract is poured onto the filter paper. The filter paper and article surface are allowed to
dry for 6-8 hours, and the filter paper is removed. The surface of the article in the beaker is then covered
with another  filter paper and the experiment is repeated with the same simulated sweat solution  four
times, for  a total  of 5 filter  paper samples. It is recommended to consider the application of pressure to
the filter paper covered article using a range of weights (i.e. one psi weight measuring 2 inches in diameter
and weighing 3.4 Ibs, or other weights consistent with typical and high-end dermal contact) in a portion
of the replicates. If the article being tested contains a barrier material  (i.e., textile covering of a couch
cushion), this  should be considered in the testing; the replicates should consider migration both with and
without the presence of such a barrier between the filter paper and the article. After collection and drying,
the five filter replicate paper samples are then extracted and analyzed for the chemical of interest.
                                         Page 39 of 41

-------
                                                              optional psi weights

                                                              filter paper

                                                              2-4 ml simulated sweat
Figure 16.  Graphic example of small-scale procedure for analyzing migration from product or
           article surface to sweat
As an alternative, a large scale experiment could be conducted to evaluate a larger surface area of an
article including up to full size (e.g., full couch cushion, pillow, and mattress). The actual surface area and
thickness of the article used in the experiment may vary but should be documented. Two filter papers
should be placed on the entire surface of the article and wetted with 25 ml of simulated sweat. Note, the
amount of simulated sweat may vary depending on the physical activity level and age of an individual so
a range of simulated sweat amounts can be considered. One psi weight should be placed on each filter.
One weight should be removed after the filter paper is thoroughly wetted; the other should be removed
six hours after application of the simulated sweat. The first situation mimics intermittent skin contact with
the article while the second mimics continuous skin contact with the article. The surface of the article is
then covered with two new dry pieces of  filter paper and the experiment is repeated with the same
simulated  sweat solution four times, for a  total of 5 tests with 10 filter paper samples collected.  Five
replicate tests are done for each sample. If the article being tested contains a barrier material (i.e., textile
covering of a couch cushion), this should be considered in the testing; the replicates  should consider
migration both with and without the presence  of such a barrier between the filter paper and the article.
The 10 filter paper samples are then extracted and analyzed for the chemical of interest.

Barrier materials and/or the surfaces of the articles themselves may or may not be treated with  various
chemicals which  are intended to promote  stain resistance, water repellence, etc. The use of materials
with these chemicals added is applicable if representative of the exposure scenario of interest. If a barrier
material of any kind is used, the experiments should be  repeated five times both with and without the
use of the barrier material.

These experiments can be described as surface migration tests which estimate the quantity of chemical
that might migrate to the skin from the surface of an article over time under certain conditions of use.
Extraction methods and analytical approaches for the skin or skin surrogate such as filter paper will vary
based on the chemical and exposure scenario of interest.

Reporting  Results and Records  Retention:
A final  report shall be  prepared,  and records  shall be retained in  accordance with 40 CFR 792,
Subpart J - Records and Reports.

For example, the following key information  should be included in the report:

    Sampling and analytical methods — description or citation, including deviation from standard
    procedure, if applicable.
    Description of simulated sweat composition (components, weight, volume)
    Description of tested  material (size, dimensions)
    Compliance with applicable human subjects research or other ethical requirements
                                         Page 40 of 41

-------
    QA/QC data: accuracy and precision of measurements

References:
Bhooshan, B., (2005). Vinylidene Chloride Testing in Mattress Barrier Samples. Tab H. Pages 535-537,
    available at http://www.cpsc.gov/PageFiles/88231/matttabh.pdf
Cobb, D. (2005). Migration of Flame retardant Chemicals in Mattress Barriers. Tab H. Pages 542-553,
    available at http://www.cpsc.gov/PageFiles/88231/matttabh.pdf
Buist, H. E., van Burgsteden, J. A., Freidig, A. P., Maas, W. J., & van de Sandt, J. J. (2010). New in vitro
    dermal absorption database and the prediction of dermal absorption under finite conditions for risk
    assessment purposes. Regulatory Toxicology and Pharmacology, 57(2), 200-209.
EPA Method—Methods for Assessing Consumer Exposure to Chemical Substances, Report EPA/747/R-
    92/003
EPA Method A Laboratory Method to Determine the Retention of Liquids on the Surface of Hands,
    Report EPA/560/585/007
Frasch, H. F., Dotson, G. S., Bunge, A. L., Chen, C. P., Cherrie, J. W., Kasting, G.  B., Kissel, J.C., Sahmel J.,
    Semple S. and Wilkinson, S. (2014). Analysis of finite dose dermal absorption data: Implications for
    dermal exposure assessment. Journal of Exposure Science and Environmental Epidemiology, 24(1),
    65-73.
Gong, M., Zhang, Y., & Weschler, C.J. (2014). Measurement of Phthalates in Skin Wipes: Estimating
    Exposure from Dermal Absorption. Environmental Science & Technology, 48, 7428-7435
Marques, M. R., Loebenberg, R., & Almukainzi, M. (2011). Simulated biological fluids with possible
    application in dissolution testing. Dissolution Technol,18(3), 15-28 (Table 17).
Ng, M. G., Semple, S., Cherrie, J. W., Christopher, Y., Northage,  C., Tielemans, E., Veroughstraete, V., &
    Van Tongeren, M. (2012). The relationship between inadvertent ingestion and dermal exposure
    pathways: A new integrated conceptual model and a database of dermal and oral transfer
    efficiencies. Annals of occupational hygiene.
Organization for Economic Co-operation and Development, 2004a. (Test Guideline 428) Skin Absorption-
    In vitro method.
Organization for Economic Co-operation and Development, 2004b. (Section 4-Other Test Guidelines)
    Guidance Document for the Conduct of Skin Absorption Studies.
Thomas, T. A., & Brundage, P. M. (2006). Quantitative Assessment of Potential Health Effects from the
    Use of Fire Retardant Chemicals in Mattresses. CPSC. Tab D. Section 2. Methods. Pages 10 -13,
    available at http://www.cpsc.gov/PageFiles/88208/matttabd.pdf
Thomas, T. A., & Brundage, P. M. (2006). Quantitative Assessment of Potential Health Effects from the
    Use of Fire Retardant Chemicals in Mattresses. CPSC. Tab D. Appendix 2. Experimental Protocol Flow
    Chart.  Pages 239-242, available at  http://www.cpsc.gov/PageFiles/88208/matttabd.pdf
Weschler, C.J., Beko, G., Koch, M., Salthammer, T., Schripp, T., Toftum, J. & Claesen, G. (2015)
    Transdermal Uptake of Diethyl Phthalate and Di(n-butyl) Phthalate Directly from Air: Experimental
    Verification. Environmental Health Perspectives,  http://dx.doi.org/10.1289/ehp.1409151.
                                         Page 41 of 41

-------