EPA/625/R-96/010b
      Compendium of Methods
      for the Determination of
     Toxic Organic Compounds
           in Ambient Air
           Second Edition
    Compendium Method TO-4A

   Determination of Pesticides and
Polychlorinated Biphenyls in Ambient
Air Using High Volume Polyurethane
 Foam (PUF) Sampling Followed by
Gas Chromatographic/Multi-Detector
         Detection (GC/MD)
      Center for Environmental Research Information
        Office of Research and Development
        U.S. Environmental Protection Agency
             Cincinnati, OH 45268

              January 1999

-------
                                       Method TO-4A
                                     Acknowledgements

This Method was prepared for publication in the Compendium of Methods for the Determination of Toxic
Organic Compounds in Ambient Air, Second Edition (EPA/625/R-96/010b), which was prepared under Contract
No. 68-C3-0315, WA No. 3-10, by Midwest Research Institute (MRI), as a subcontractor to Eastern Research
Group, Inc. (ERG), and under the sponsorship of the U.S. Environmental Protection Agency (EPA). Justice A.
Manning, John O. Burckle, and Scott R. Hedges, Center for Environmental Research Information (CERI), and
Frank F. McElroy, National Exposure Research Laboratory (NERL), all in the EPA Office of Research and
Development (ORD), were responsible for overseeing the preparation of this method. Additional support was
provided by other members of the  Compendia Workgroup, which include:
         John O. Burckle, U.S. EPA, ORD, Cincinnati, OH
         James L. Cheney, Corps of Engineers, Omaha, NB
         Michael Davis, U.S. EPA, Region 7, KC, KS
         Joseph B. Elkins Jr., U.S. EPA, OAQPS, RTP, NC
         Robert G. Lewis, U. S. EPA, NERL, RTP, NC
         Justice A. Manning, U.S. EPA, ORD, Cincinnati, OH
     •    William A. McClenny, U.S. EPA, NERL, RTP, NC
         Frank F. McElroy, U.S.  EPA, NERL, RTP, NC
         Heidi Schultz, ERG, Lexington, MA
     •    William T.  "Jerry" Winberry, Jr., EnviroTech Solutions, Gary, NC
Method TO-4 was originally published  in April of  1984 as one of a series of peer reviewed  methods in
"Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air, " EPA 600/4-
89-018.  In an effort to keep these methods consistent with current technology, Method TO-4 has been revised
and updated as Method TO-4A in this Compendium to incorporate new or improved sampling and analytical
technologies. In addition, this method incorporates ASTM Method D 4861-94, Standard Practice for Sampling
and Analysis of Pesticides and Poly chlorinated Biphenyls in Air.

This Method is the result of the efforts of many individuals.  Gratitude goes to each person  involved in the
preparation and review of this methodology.

Author(s)
     •    William T.  "Jerry" Winberry, Jr., EnviroTech Solutions, Gary, NC
         Ralph Riggin, Battelle Laboratories, Columbus, OH
         Robert G. Lewis, U. S. EPA, NERL, RTP, NC
Peer Reviewers
         Irene D. DeGraff, Supelco, Bellefonte, PA
     •    Ronald A. Hiles, Indiana University, Bloomington, IN
         Lauren Drees, U.S. EPA, NRMRL, Cincinnati, OH
Finally, recognition is given to Frances Beyer, Lynn Kaufman, Debbie Bond, Cathy Whitaker, and Kathy Johnson
of Midwest Research Institute's Administrative  Services staff whose dedication and persistence during the
development of this manuscript has enabled it's production.

                                        DISCLAIMER

This Compendium has been subjected to the Agency's peer and administrative review, and it has been
approved for publication as  an EPA document. Mention of trade names or commercial products does
not constitute endorsement or recommendation for use.

-------
                                     METHOD TO-4A

                               Determination of Pesticides and
                            Polychlorinated Biphenyls in Ambient
                            Air Using High Volume Polyurethane
                              Foam (PUF) Sampling Followed by
                            Gas Chromatographic/Multi-Detector
                                     Detection (GC/MD)

                                  TABLE OF CONTENTS

                                                                                       Page


1.  Scope    	     4A-1

2.  Summary of Method	     4A-1

3.  Significance	     4A-2

4.  Applicable Documents	     4A-2
     4.1  ASTM Standards  	     4A-2
     4.2  EPA Documents	     4A-2
     4.3  Other Documents	     4A-3

5.  Definitions 	     4A-3

6.  Interferences	     4A-4

7.  Safety    	     4A-4

8.  Apparatus	     4A-5
     8.1  Sampling	     4A-5
     8.2  Sample Clean-up and Concentration	     4A-6
     8.3  Sample Analysis	     4A-7

9.  Equipment and Materials	     4A-7
     9.1  Materials for Sample Collection  	     4A-7
     9.2  Sample Extraction and Concentration	     4A-8
     9.3  GC/MS Sample Analysis	     4A-8

10. Preparation of PUF Sampling Cartridge 	     4A-9
     10.1 Summary of Method	     4A-9
     10.2 Preparation of Sampling Cartridge  	     4A-9
     10.3 Procedure for Certification of PUF Cartridge Assembly	    4A-10

11. Assembly, Calibration and Collection Using High-Volume Sampling System  	    4A-11
     11.1 Description of Sampling Apparatus	    4A-11
     11.2 Calibration of Sampling System 	    4A-11
     11.3 Sample  Collection 	    4A-18
                                            in

-------
                             TABLE OF CONTENTS (continued)

                                                                                         Page
12.  Sample Extraction Procedure	    4A-20
     12.1 Sample Extraction 	    4A-20
     12.2 Sample Cleanup	    4A-21

13.   Analytical Procedure	    4A-22
     13.1 Analysis Organochlorine Pesticides by Capillary Gas Chromatography
          with Electron Capture Detector (GC/ECD)  	    4A-22
     13.2 Analysis of Organophosphorus Pesticides by Capillary Gas
          Chromatography with Flame Photometric or Nitrogen-Phosphorus
          Detectors (GC/FPD/NPD)	    4A-23
     13.3 Analysis of Carbamate and Urea Pesticides by Capillary Gas
          Chromatography with Nitrogen-Phosphorus Detector	    4A-23
     13.4 Analysis of Carbamate, Urea, Pyrethroid, and Phenolic Pesticides by
          High Performance Liquid Chromatography (HPLC) 	    4A-24
     13.5 Analysis of Pesticides and PCBs by Gas Chromatography with Mass
          Spectrometry Detection (GC/MS)	    4A-24
     13.6 Sample Concentration	    4A-25

14.   Calculations	    4A-25
     14.1 Determination of Concentration  	    4A-25
     14.2 Equations	    4A-25

15.   Performance Criteria and Quality Assurance  	    4A-27
     15.1 Standard Operating Procedures (SOPs) 	    4A-28
     15.2 Process, Field, and Solvent Blanks	    4A-28
     15.3 Method Precision and Bias  	    4A-28
     15.4 Method Safety	    4A-29

16.   References	    4A-29
                                             IV

-------
                                       METHOD TO-4A

                   Determination of Pesticides and Polychlorinated Biphenyls in
               Ambient Air Using High Volume Polyurethane Foam (PUF) Sampling
                        Followed by Gas Chromatographic/Multi-Detector
                                      Detection (GC/MD)
1. Scope
1.1 This document describes a method for sampling and analysis of a variety of common pesticides and for
polychlorinated biphenyls (PCBs) in ambient air. The procedure is based on the adsorption of chemicals from
ambient air on polyurethane foam (PUF) using a high volume sampler.

1.2 The high volume PUF sampling procedure is applicable to multicomponent atmospheres containing common
pesticide concentrations from 0.001 to 50 ,ug/m3 over 4- to 24-hour sampling periods. The limits of detection
will depend on the nature of the analyte and the length of the sampling period.

1.3  Specific compounds for which the method has been employed are listed in Table  1.  The analytical
methodology described in Compendium Method TO-4A is currently employed by laboratories throughout the
U.S. The sampling methodology has been formulated to meet the needs of common pesticide and PCB sampling
in ambient air.

1.4 Compendium Method TO-4 was originally published in 1989 (1).  Further updates of the sampling protocol
were published as part of Compendium Method TO-13 (2).  The method was further modified for indoor air
application in  1990  (3).  In an effort to keep the method consistent with current technology, Compendium
Method TO-4 has incorporated the sampling and analytical procedures in ASTM Method D4861-94 (4) and is
published here as Compendium Method TO-4A.
2.  Summary of Method

2.1 A high-volume (~8 cfm) sampler is used to collect common pesticides and PCBs on a sorbent cartridge
containing PUF. Airborne particles may also be collected, but the sampling efficiency is not known (5).  The
sampler is operated for 24-hours, after which the sorbent is returned to the laboratory for analysis.

2.2 Pesticides and PCBs are extracted from the sorbent cartridge with 10 percent diethyl ether in hexane and
determined by gas chromatography coupled with an electron capture detector (BCD), nitrogen-phosphorus
detector (NPD), flame photometric detector (FPD), Hall electrolytic conductivity detector (HECD), or a mass
spectrometer (MS). For common pesticides, high performance liquid chromatography (HPLC) coupled with an
ultraviolet (UV) detector or electrochemical detector may be preferable.

2.3 Interferences resulting from analytes having similar retention times during GC analysis are resolved by
improving the resolution or separation, such as by changing the chromatographic column or operating parameters,
or by fractionating the sample by column chromatography.
January 1999        Compendium of Methods for Toxic Organic Air Pollutants            Page 4A-1

-------
Method TO-4A                                                                  Pesticides/PCBs
3.  Significance

3.1 Pesticide usage and environmental distribution are common to rural and urban areas of the United States.
The application of pesticides can cause adverse health effects to humans by contaminating soil, water, air, plants,
and animal life.  PCBs are less widely used, due to extensive restrictions placed on their manufacturer. However,
human exposure to PCBs continues to be a problem because of their presence in various electrical products.

3.2 Many pesticides and PCBs exhibit bioaccumulative, chronic health effects; therefore, monitoring the presence
of these compounds in ambient air is of great importance.

3.3 The relatively low levels of such compounds in the environment requires the use of high volume sampling
techniques to acquire sufficient  sample for analysis.  However, the volatility of these compounds prevents
efficient collection on filter media.  Consequently, Compendium Method TO-4A utilizes both a filter and a PUF
backup cartridge which provides for efficient collection of most common pesticides, PCBs, and many other
organics within the same volatility range.

3.4 Moreover, modifications to this method has been successfully applied to measurement of common pesticides
and PCBs in outdoor air (6), indoor air (3) and for personal respiratory exposure monitoring (3).
4.  Applicable Documents

4.1 ASTM Standards

   • D1356 Definition of Terms Relating to Atmospheric Sampling and Analysis
   • D4861-94 Standard Practice for Sampling and Analysis of Pesticides and Poly chlorinate dBiphenyls
     in Air
   • E260 Recommended Practice for General Gas Chromatography Procedures
   • E355 Practice for Gas Chromatography Terms and Relationships
   • D3686 Practice for Sampling Atmospheres to Collect Organic Compound Vapors (Activated Charcoal
     Tube Adsorption Method
   • D3687 Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube
     Adsorption
   • D4185  Practice for  Measurement of Metals in  Workplace Atmosphere  by Atomic Absorption
     Spectrophotometry

4.2 EPA Documents

   • Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air: Method
     TO-10, Second Supplement, U. S. Environmental Protection Agency, EPA 600/4-89-018, March 1989.
   • Manual of Analytical Methods for Determination of Pesticides  in Humans and Environmental
     Standards, U. S.  Environmental Protection Agency, EPA 600/8-80-038, June 1980.
   • Compendium of Methods for the Determination of Air Pollutants in Indoor Air: Method IP-8, U. S.
     Environmental Protection Agency, EPA 600/4-90-010, May 1990.
Page 4A-2            Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

4.3 Other Documents

   •  Code of Federal Regulations, Title 40, Part 136, Method 604


5. Definitions

[Note: Definitions used in this document and in any user-prepared Standard operating procedures (SOPs)
should be consistent with ASTMD1356, E260, and E355.  All abbreviations and symbols are defined within
this document at point of use.]

5.1 Sampling efficiency (SE)-ability of the sampling medium to trap analytes of interest. The percentage of
the analyte of interest collected and retained by the sampling medium when it is introduced as a vapor in air or
nitrogen into the air sampler and the sampler is operated under normal conditions for a period of time equal to
or greater than that required for the intended use is indicated by %SE.

5.2 Retention efficiency (RE)-ability of sampling medium to retain a compound added (spiked) to it in liquid
solution.

5.3 Retention time (RT)-time to elute a specific chemical from a chromatographic column, for a specific carrier
gas flow rate, measured from the time the chemical is injected into the gas stream until it appears at the detector.

5.4 Relative retention time (RRT)-a rate of RTs for two chemicals for the same chromatographic column and
carrier gas flow  rate, where the denominator represents a reference  chemical.

5.5  Method detection limit (MDL)-the minimum concentration of a substance that can be measured and
reported with confidence and that the value is above zero.

5.6 Kuderna-Danish apparatus-the Kuderna-Danish (K-D) apparatus is a system for concentrating materials
dissolved in volatile solvents.

5.7 MS-SIM-the GC is coupled to a mass spectrometer where the instrument is programmed to acquire data for
only the target compounds and to disregard all others, thus operating in the select ion monitoring mode (SIM).
This is performed using SIM coupled to retention time discriminators.  The SIM analysis procedure provides
quantitative results.

5.8 Sublimation-the direct passage of a substance from the  solid  state to the gaseous state and back into the
solid form without any time appearing in the liquid state. Also applied to the conversion of solid to vapor without
the later return to solid state, and to a conversion directly from the  vapor phase to the solid state.

5.9 Surrogate standard-a chemically compound (not expected to occur in the environmental sample) which is
added to each sample, blank and matrix spiked sample before extraction and analysis.  The recovery of the
surrogate  standard is used to monitor unusual matrix effects, gross sample processing errors, etc.  Surrogate
recovery is evaluated for acceptance by determining whether the measured concentration falls within acceptable
limits.
January 1999        Compendium of Methods for Toxic Organic Air Pollutants            Page 4A-3

-------
Method TO-4A                                                                    Pesticides/PCBs
6. Interferences

6.1 Any gas or liquid chromatographic separation of complex mixtures of organic chemicals is subject to serious
interference problems due to coelution of two or more compounds.  The use of capillary or microbore columns
with superior resolution or two or more columns of different polarity will frequently eliminate these problems.
In addition, selectivity may be further enhanced by use of a MS operated in the selected ion monitoring (SIM)
mode as the GC detector.  In this mode, co-eluting compounds can often be determined.

6.2  The BCD responds to a wide variety of organic compounds.  It is likely that such compounds will be
encountered as interferences during GC/ECD analysis. The NPD, FPD, and HECD detectors are  element specific,
but are still subject to interferences.  UV detectors for HPLC are nearly universal, and the electrochemical
detector may also respond to a variety of chemicals. Mass spectrometric analyses will generally provide positive
identification of specific compounds.

6.3 PCBs and certain common pesticides (e.g., chlordane) are complex mixtures of individual compounds which
can cause difficulty in accurately quantifying a particular formulation in a multiple component mixture. PCBs
may interfere with the determination of pesticides.

6.4 Contamination of glassware and sampling apparatus with traces of pesticides or PCBs can be a major source
of error, particularly at lower analyte concentrations.  Careful attention to cleaning and handling procedures is
required during all steps of sampling and analysis to minimize this source of error.

6.5 The general  approaches listed below should be followed to minimize interferences.

    6.5.1  Polar compounds, including certain pesticides (e.g., organophosphorus and carbamate classes) can be
removed by column chromatography on alumina.  Alumina  clean-up will permit analysis of most common
pesticides and PCBs (7).
    6.5.2 PCBs may be separated from other common pesticides by column chromatography on silicic acid (8,9).
    6.5.3 Many pesticides can be fractionated into groups by column chromatography on Florisil (9).
7. Safety

7.1 The toxicity or carcinogencity of each reagent used in this method has not been precisely defined; however,
each chemical compound should be treated as a potential health hazard.  From this viewpoint, exposure to these
chemicals must be reduced to  the lowest possible level by whatever means available.  The laboratory is
responsible for maintaining a current awareness file of Occupational Safety and Health Administration (OSHA)
regulations regarding the safe handling of the chemicals specified in this method. A reference file of material data
handling sheets should also be made  available to all personnel involved in the chemical analysis.  Additional
references to laboratory safety are available and have been identified for the analyst (10-12).

7.2 PCBs have been classified as a known or suspected, human or mammalian carcinogen. Many of the other
common pesticides  have been classified as carcinogens.  Care must be exercised when working with these
substances.  This method does  not purport to address all safety problems associated with  its use.  It is the
responsibility of whoever uses this method to consult and establish appropriate safety and health practices and
Page 4A-4            Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

determine the applicability of regulatory limitations prior to use.  The user should be thoroughly familiar with
the chemical and physical properties of targeted substances.

7.3 Treat all target analytes as carcinogens. Neat compounds should be weighed in a glove box. Spent samples
and unused standards are toxic waste and should be disposed according to regulations. Regularly check counter
tops and equipment with "black light" for fluorescence as an indicator of contamination.

7.4 The collection efficiency for common pesticides and PCBs has been demonstrated to be greater than 95
percent for the sampling configuration described in the method (filter and backup adsorbent). Therefore, no field
recovery evaluation will occur as part of this procedure.
8. Apparatus

[Note: This method was developed using the PS-1 semi-volatile sampler provided by General Metal Works,
Village of Cleves, OH as a guideline.  EPA has experience in use of this equipment during various field
monitoring programs  over the last several years.  Other manufacturers' equipment should work as well.
However, modifications to these procedures may be necessary if another commercially available sampler is
selected.]

8.1 Sampling

    8.1.1 High-volume sampler (see Figure 1).  Capable of pulling ambient air through the filter/adsorbent
cartridge at a flow rate of approximately 8 standard cubic feet per minute (scfm) (0.225 std mVmin) to obtain
a total sample volume of greater than 300 scm over a 24-hour period. Major manufacturers are:

    •  Tisch Environmental, Village of Cleves, OH
    •  Andersen Instruments Inc., 500 Technology Ct, Smyrna, GA
    •  Thermo Environmental Instruments, Inc., 8 West Forge Parkway, Franklin, MA

    8.1.2 Sampling module (see Figure 2). Metal filter holder (Part 2) capable of holding a 102-mm circular
particle filter supported by a 16-mesh stainless-steel screen and attaching to a metal cylinder (Part 1) capable of
holding a 65-mm O.D. (60-mm I.D.) x 125-mm borosilicate glass sorbent cartridge containing PUF. The filter
holder is equipped with inert sealing gaskets (e.g., polytetrafluorethylene) placed on either side of the filter.
Likewise, inert, pliable gaskets (e.g., silicone rubber) are used to provide an air-tight seal at each end of the glass
sorbent cartridge. The glass sorbent cartridge is indented 20 mm from the lower end to provide a support for a
16-mesh stainless-steel screen that holds the sorbent.  The glass sorbent cartridge fits into Part 1, which is
screwed onto Part 2 until the sorbent cartridge is sealed between the silicone gaskets. Major manufacturers are:

    •  Tisch Environmental, Village of Cleves, OH
    •  Andersen Instruments Inc., 500 Technology Ct., Smyrna, GA
    •  Thermo Environmental Instruments, Inc., 8 West Forge Parkway, Franklin, MA

A field portable unit has been developed by EPA (see Figure 3).
    8.1.3 High-volume sampler calibrator.  Capable of providing multipoint resistance for the high-volume
sampler. Major manufacturers are:
January 1999         Compendium of Methods for Toxic Organic Air Pollutants            Page 4A-5

-------
Method TO-4A	Pesticides/PCBs

    •  Tisch Environmental, Village of Cleves, OH
    •  Andersen Instruments Inc., 500 Technology Ct, Smyrna, GA
    •  Thermo Environmental Instruments, Inc., 8 West Forge Parkway, Franklin, MA

    8.1.4 Ice chest. To hold samples at <4°C or below during shipment to the laboratory after collection.
    8.1.5 Data sheets. For each sample for recording the location and sample time, duration of sample, starting
time,  and volume of air sampled.

8.2 Sample Clean-up and Concentration (see Figure 4).

    8.2.1 Soxhlet apparatus extractor (see Figure 4a). Capable of extracting filter and adsorbent cartridges
(2.3"  x 5" length), 1,000 mL flask, and condenser, best source.
    8.2.2 Pyrex glass tube furnace system. For activating silica gel at 180°C under purified nitrogen gas purge
for an hour, with capability of raising temperature gradually, best source.
    8.2.3 Glass vial.  40 mL, best source.
    8.2.4 Erlenmeyer flask. 50 mL, best source.

[Note: Reuse of glassware should be minimized to avoid the risk of cross contamination. All glassware that
is used, especially glassware that is reused, must be scrupulously cleaned as soon as possible after use. Rinse
glassware with the last solvent used in it and then with high-purity acetone and hexane.  Wash  with hot water
containing detergent. Rinse with copious amount of tap water and several portions of distilled water. Drain,
dry, and heat in a muffle furnace at 400° C for 4 hours.  Volumetric glassware must not be heated in a muffle
furnace; rather, it should be rinsed with high-purity acetone and hexane.  After the glassware is dry and cool,
rinse it with hexane, and  store it inverted or capped with solvent-rinsed aluminum foil in a clean
environment.]

    8.2.5 White cotton gloves. For handling cartridges and filters, best source.
    8.2.6 Minivials. 2 mL, borosilicate glass, with conical reservoir and screw caps lined with Teflon®-faced
silicone disks, and a vial holder, best source.
    8.2.7 Teflon®-coated stainless steel spatulas and spoons. Best source.
    8.2.8 Kuderna-Danish (K-D) apparatus (see Figure 4b). 500 mL evaporation flask (Kontes K-570001-
500 or equivalent), 10 mL graduated concentrator tubes (Kontes K570050-1025 or equivalent) with ground-glass
stoppers, and 3-ball macro Snyder Column (Kontes K-570010500,  K-50300-0121, and K-569001-219, or
equivalent), best source.
    8.2.9 Adsorption column for column chromatography (see Figure 4c).  1-cm x 10-cm with stands.
    8.2.10 Glove box. For working with extremely toxic standards and reagents with explosion-proof hood for
venting fumes from solvents, reagents, etc.
    8.2.11 Vacuum oven. Vacuum drying oven system capable of maintaining a vacuum at 240 torr (flushed
with nitrogen) overnight.
    8.2.12 Concentrator tubes and a nitrogen evaporation apparatus with variable flow rate. Best source.
    8.2.13  Laboratory refrigerator. Best source.
    8.2.14  Boiling chips. Solvent extracted, 10/40 mesh silicon carbide or equivalent, best source.
    8.2.15 Water bath. Heated, with concentric ring cover, capable of ±5°C temperature control, best source.
    8.2.16  Nitrogen evaporation apparatus.  Best source.
    8.2.17  Glass wool. High purity grade, best source.
Page 4A-6            Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

8.3 Sample Analysis

   8.3.1 Gas chromatograph (GC). The GC system should be equipped with appropriate detector(s) and
either an isothermally controlled or temperature programmed heating oven. Improved detection limits may be
obtained with a GC equipped with a cool on-column or splitless injector.
   8.3.2 Gas chromatographic column. As an example, a 0.32-mm (I.D.) x 3-mm DB-5, DB-17, DB-608,
DB-1701 are available.  Other columns may also provide acceptable results.
   8.3.3 HPLC column. As an example, a 4.6-mm x 25-cm Zorbax SIL or uBondpak C-18. Other columns
may also provide acceptable results.
   8.3.4 Microsyringes. 5 uL volume or other appropriate sizes.
   8.3.5 Balance. Mettler balance or equivalent.
   8.3.6 All required syringes, gases, and other pertinent supplies. To operate the GC/MS system.
   8.3.7 Pipettes, micropipettes, syringes, burets, etc.  To make calibration and spiking solutions, dilute
samples if necessary, etc., including syringes for accurately measuring volumes such as 25 uL and 100 uL.
9. Equipment and Materials

9.1 Materials for Sample Collection (see Figure 5)

   9.1.1  Quartz fiber filter.  102-millimeter bindless quartz microfiber filter, Whatman Inc., 6 Just Road,
Fairfield, NJ 07004, Filter Type QMA-4.
   9.1.2 Polyurethane foam (PUF) plugs (see Figure 5a). 3-inch thick sheet stock polyurethane type (density
.022 g/cm3). The PUF should be of the polyether type used for furniture upholstery, pillows, and mattresses. The
PUF cylinders (plugs) should be slightly larger in diameter than the internal diameter of the cartridge. Sources
of equipment are Tisch Environmental, Village of Cleves, OH; University Research Glassware, 116 S. Merritt
Mill Road, Chapel Hill, NC; Thermo Environmental Instruments, Inc., 8 West Forge Parkway, Franklin, MA;
Supelco, Supelco Park, Bellefonte, PA; and SKC Inc., 334 Valley View Road, Eighty Four, PA.
   9.1.3  Teflon® end  caps (see Figure  5a).  For sample cartridge.  Sources of equipment are  Tisch
Environmental, Village of Cleves, OH and University Research Glassware, Chapel Hill, NC.
   9.1.4 Sample cartridge aluminum shipping containers (see Figure 5b). For sample cartridge shipping.
Sources of equipment are Tisch Environmental, Village of Cleves, OH and University Research Glassware,
Chapel Hill, NC.
   9.1.5  Glass sample cartridge (see Figure 5a).  For sample collection. Sources of equipment are  Tisch
Environmental, Village of Cleves, OH;  Thermo Environmental Instruments, Inc., 8 West Forge Parkway,
Franklin, MA; University Research Glassware, 116 S. Merritt Mill Road, Chapel Hill, NC; and Supelco, Supelco
Park, Bellefonte, PA.
   9.1.6 Aluminum foil. Best source.
   9.1.7 Hexane, reagent grade. Best source.

9.2 Sample Extraction and Concentration

   9.2.1 Methylene chloride. Chromatographic grade, glass-distilled, best source.
   9.2.2 Sodium sulfate-anhydrous (ACS). Granular (purified by washing with methylene chloride followed
by heating at 400°C for 4 hours in a shallow tray).
   9.2.3 Boiling chips. Solvent extracted or heated in a muffle furnace at 450°C for 2 hours, approximately
10/40 mesh (silicon carbide or equivalent).
January 1999        Compendium of Methods for Toxic Organic Air Pollutants            Page 4A-7

-------
Method TO-4A	Pesticides/PCBs

   9.2.4 Nitrogen. High purity grade, best source.
   9.2.5 Ether. Chromatographic grade, glass-distilled, best source.
   9.2.6 Hexane. Chromatographic grade, glass-distilled, best source.
   9.2.7 Dibromobiphenyl.  Chromatographic grade, best source. Used for internal standard.
   9.2.8 Decafluorobiphenyl. Chromatographic grade, best source. Used for internal standard.
   9.2.9 Glass wool. Silanized, extracted with methylene chloride and hexane, and dried.
   9.2.10 Diethyl ether. High purity, glass distilled.
   9.2.11 Hexane. High purity, glass distilled.
   9.2.12 Silica gel. High purity, type 60, 70-230 mesh.
   9.2.13 Round bottom evaporative flask.  500 mL, 1 24/40 joints, best source.
   9.2.14 Capacity soxhlet extractors. 500 mL, with reflux condensers, best source.
   9.2.15 Kuderna-Danish concentrator. 500 mL, with Snyder columns, best source.
   9.2.16 Graduated concentrator tubes.  10 mL, with 19/22 stoppers, best source.
   9.2.17 Graduated concentrator tubes.  1 mL, with 14/20 stoppers,  best source.
   9.2.18 TFE fluorocarbon tape. 1/2 in., best source.
   9.2.19 Filter tubes.  Size 40-mm (I.D.) x 80-mm.
   9.2.20 Serum vials.  1 mL and 5 mL, fitted with caps lined with TFE fluorocarbon.
   9.2.21 Pasteur pipetter.  9 in., best source.
   9.2.22 Glass wool. Fired at 500°C, best source.
   9.2.23 Alumina. Activity Grade IV, 100/200 mesh.
   9.2.24 Glass Chromatographic column.  2-mm I.D. x 15-cm long.
   9.2.25 Vacuum oven. Connected to water aspirator, best source.
   9.2.26 Die. Best source.
   9.2.27 Ice chest. Best source.
   9.2.28 Silicic Acid. Pesticide quality, best source.
   9.2.29 Octachloronaphthalene (OCN). Research grade, best source.
   9.2.30 Florisil.  Pesticide quality, best source.

9.3 GC Sample Analysis

   9.3.1  Gas cylinders of hydrogen, nitrogen, argon/methane, and helium.  Ultra high purity, best source.
   9.3.2 Combustion air. Ultra high purity, best source.
   9.3.3  Zero air.  Zero air may be obtained from a cylinder or zero-grade  compressed air scrubbed with
Drierite® or silica gel and 5A molecular sieve or activated charcoal, or by catalytic cleanup of ambient air. All
zero air should be passed through a liquid argon cold trap for final cleanup.
   9.3.4 Chromatographic-grade stainless steel tubing and stainless steel fitting.  For interconnections,
Alltech Applied Science, 2051 Waukegan Road, Deerfield, IL 60015, 312-948-8600, or equivalent.

[Note: All such materials in contact with the sample, analyte,  or support gases prior to analysis should be
stainless steel or other inert metal. Do not use plastic or Teflon® tubing or fittings.]
10. Preparation of PUF Sampling Cartridge

[Note:  This method was developed using the PS-1 sample cartridge provider by General Metal Works, Village
ofCleves, OH as a guideline.  EPA has experience in use of this equipment during various field monitoring
Page 4A-8           Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

programs over the last several years.  Other manufacturers' equipment should work as well.  However,
modifications to these procedures may be necessary if another commercially available sampler is selected.]

10.1  Summary of Method

    10.1.1 This part of Compendium Method TO-4A discusses pertinent information regarding the preparation
and cleaning of the filter, adsorbent, and filter/adsorbent cartridge assembly. The separate batches of filters and
adsorbents are extracted with the appropriate solvent.
    10.1.2 At least one PUF cartridge assembly and one filter from each batch, or 10 percent of the batch,
whichever is greater, should be tested and certified clean before the batch is considered for field use.

10.2  Preparation of Sampling Cartridge

    10.2.1 Bake the Whatman QMA-4 quartz filters at 400°C for 5 hours before use.
    10.2.2 Set aside the filters in a clean container for shipment to the field or prior to combining with the PUF
glass cartridge assembly for certification prior to field deployment.
    10.2.3 The PUF plugs are 6.0-cm diameter cylindrical plugs cut from 3-inch sheet stock and should fit, with
slight compression, in the glass cartridge, supported by the wire screen (see Figure 2).  During cutting, rotate the
die at high speed (e.g., in a drill press) and continuously lubricate with deionized or distilled water.  Pre-cleaned
PUF plugs can be obtained from many of the commercial sources identified in Section 9.1.2.
    10.2.4 For initial cleanup, place the PUF plugs in a Soxhlet apparatus and extract with acetone for 16 hours
at approximately  4 cycles per hour.  When cartridges are reused, use  diethyl ether/hexane  (10 percent
volume/volume [v/v]) as the cleanup solvent.

[Note: A modified PUF cleanup procedure can be used to remove unknown interference components of the
PUF blank.  This method consists of rinsing 50 times with toluene, acetone, and diethyl ether/hexane (5 to
10 percent v/v), followed by Soxhlet extraction. The extracted PUF is placed in a vacuum oven connected to
a water aspirator and dried at room temperature for approximately 2 to 4 hours (until no solvent odor is
detected). Alternatively, they may be dried at room temperature in an air-tight container with circulating
nitrogen (zero grade).  Place the clean PUF plug into a labeled glass sampling cartridge using gloves and
forceps.  Wrap the cartridge  with hexane-rinsed aluminum foil and placed in a jar fitted with TFE
fluorocarbon-linedcaps.  The foil wrapping may also be marked for identification using a blunt probe.  The
extract from  the  Soxhlet extraction procedure from each batch may be analyzed  to determine  initial
cleanliness prior to certification.]

    10.2.5 Fit a nickel or stainless steel screen (mesh size 200/200) to the bottom of a hexane-rinsed glass
sampling cartridge to retain the PUF adsorbents, as illustrated in Figure 2. Place the Soxhlet-extracted, vacuum-
dried PUF (2.5-cm thick by 6.5-cm diameter) on top of the screen in the glass sampling cartridge using polyester
gloves.
    10.2.6 Wrap the sampling cartridge with hexane-rinsed aluminum foil, cap with the Teflon® end caps, place
in a cleaned labeled aluminum shipping container, and seal with Teflon® tape.  Analyze at least 1 PUF plug from
each batch of PUF plugs using the procedure described in Section 10.3, before the batch is considered acceptable
for field use.  A blank level of <10 ng/plug and filter for single component compounds is considered to be
acceptable.  For multiple component mixtures (e.g., PCBs), the blank level should be <100 ng/plug and filter.
Cartridges are considered clean for up to 30 days from date of certification when stored in their sealed containers.
January 1999         Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-9

-------
Method TO-4A	Pesticides/PCBs

10.3 Procedure for Certification of PUF Cartridge Assembly

    10.3.1 Extract 1 filter and PUF adsorbent cartridge by Soxhlet extraction and concentrate using a Kuderna-
Danish (K-D) evaporator for each lot of filters and cartridges sent to the field.
    10.3.2 Assemble the Soxhlet apparatus. Charge the  Soxhlet apparatus (see Figure 4a) with 300 mL of the
extraction solvent [10 percent (v/v) diethyl ether/hexane] and reflux for 2 hours.  Let the apparatus cool,
disassemble it, and discard the used extraction solvent.  Transfer the filter and PUF glass cartridge to the Soxhlet
apparatus (the use of an extraction thimble is optional).

[Note:  The filter and adsorbent assembly are extracted together in order to reach  detection  limits, to
minimize cost and to prevent misinterpretation of the data. Separate analyses of the filter and P UF would not
yield useful information about the physical state of most of the common pesticides andPCBs at the time of
sampling due to evaporative losses of the analyte from the filter during sampling.]

    10.3.3  Add between 300 and 350 mL of diethyl ether/hexane (10 percent v/v) to the Soxhlet apparatus.
Reflux the sample for 18 hours at a rate of at least 3 cycles per hour. Allow to cool, then disassemble the
apparatus.
    10.3.4 Assemble a K-D concentrator (see Figure 4b) by attaching a 10-mL concentrator tube to a 500-mL
evaporative flask.
    10.3.5  Transfer the extract by pouring it through a  drying column containing about 10 cm of anhydrous
granular sodium sulfate (see Figure 4c) and collect the extract in the K-D concentrator.  Rinse the Erlenmeyer
flask and column with 20 to 30 mL of 10 percent diethyl ether/hexane to complete the quantitative transfer.
    10.3.6 Add 1 or 2 clean boiling chips and attach a 3-ball Snyder column to the evaporative flask. Pre-wet
the Snyder column by adding about 1 mL of the extraction solvent to the top of the column. Place the K-D
apparatus on a hot water bath (50°C) so that the concentrator tube is partially immersed in the hot water, and the
entire lower rounded surface of the flask is bathed with hot vapor. Adjust the vertical position of the  apparatus
and the water temperature as required to complete the concentration in one hour. At the proper rate of distillation,
the balls of the column will actively chatter but the chambers will not flood with condensed solvent. When the
apparent volume of liquid reaches approximately 5 mL, remove the K-D apparatus from the water bath and allow
it to drain and cool for at least 5 minutes.  Remove the Snyder column and rinse the flask and its lower joint into
the concentrator tube with 5 mL of hexane. A 5-mL syringe is recommended for this operation.

[Note: The solvent may have  to be exchanged to another solvent to meet the requirements of the analytical
procedure selected for the target analytes.]

    10.3.7 Concentrate the extract to 1 mL and analyze  according to Section 13.
    10.3.8 Acceptable levels of common pesticides must be less than 10 ng for each pair of filter and adsorbent
assembly analyzed. For multiple component mixtures (e.g., PCBs), the blank level should be less than 100 ng
for each pair of filter and adsorbent. Once certified clean, the cartridges can be shipped to the field without being
chilled.
11. Assembly, Calibration and Collection Using High-Volume Sampling System

[Note: This method was developed using the PS-1 semi-volatile sampler provided by General Metal Works,
Village ofCleves,  OH as a guideline.  EPA has experience in use of this equipment during various field
monitoring programs over the last several years.  Other manufacturers' equipment should work as well.
Page 4A-10           Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

However, modifications to these procedures may be necessary if another commercially available sampler is
selected.]

11.1 Description of Sampling Apparatus

The entire sampling system is diagrammed in Figure 1. This apparatus was developed to operate at a rate of 4
to 10 scfm (0.114 to 0.285 std mVmin) and is used by EPA for high-volume sampling of ambient air.  The
method write-up presents the use of this device.

The sampling module (see Figure 2) consists of a filter and a glass sampling cartridge containing the PUF utilized
to concentrate common pesticides and PCBs from the air. A field portable unit has been developed by EPA (see
Figure 3).

11.2 Calibration of Sampling System

Each sampler should be calibrated (1) when new, (2) after major repairs or maintenance, (3) whenever any audit
point deviates from the calibration curve by more than  7 percent, (4)  before/after each sampling event, and
(5) when a different sample collection media, other than that which the sampler was originally calibrated to, will
be used for sampling.
    11.2.1 Calibration of Orifice Transfer Standard. Calibrate the modified high volume air sampler in the
field using a calibrated orifice flow rate transfer standard.  Certify the orifice transfer standard in the laboratory
against a positive displacement rootsmeter (see Figure 6).  Once certified, the recertification is performed rather
infrequently if the orifice is protected from damage. Recertify the orifice transfer standard performed once per
year utilizing a set of five multiple resistance plates.

[Note:  The set of five multihole resistance plates are  used to change the flow through the orifice so that
several points can be obtained for the orifice calibration curve. The following procedure outlines the steps
to calibrate the orifice transfer standard in the laboratory.]

      11.2.1.1  Record the room temperature (Tj in °C) and barometric pressure (Pb in mm Hg) on the Orifice
Calibration Data Sheet (see Figure 7). Calculate the room temperature in K (absolute temperature) and record
on Orifice Calibration Data Sheet.

                                      T1inK = 273°+T1in°C

      11.2.1.2  Set up laboratory orifice calibration equipment as illustrated in Figure 6. Check the oil level of
the rootsmeter prior to starting.  There are 3 oil level indicators,  1 at the clear plastic end and 2 site glasses, 1 at
each end of the measuring chamber.
      11.2.1.3 Check for leaks by clamping both manometer lines, blocking the orifice with cellophane tape,
turning on the high volume motor, and noting any change in the rootsmeter's reading. If the rootsmeter's reading
changes, there is a leak in the system.  Eliminate the leak before proceeding. If the rootsmeter's reading remains
constant, turn off the hi-vol motor, remove the cellophane tape, and unclamp both manometer lines.
      11.2.1.4 Install the 5-hole resistance plate between the orifice and the filter adapter.
      11.2.1.5  Turn manometer tubing connectors 1 turn counter-clockwise. Make sure all connectors are open.
      11.2.1.6  Adjust both manometer midpoints by sliding their movable scales until the zero point corresponds
with the meniscus. Gently shake or tap to remove any air bubbles and/or liquid remaining on tubing connectors.
(If additional liquid is required for the water manometer, remove tubing connector and add clean water.)
January 1999         Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-11

-------
Method TO-4A	Pesticides/PCBs

      11.2.1.7 Turn on the high volume motor and let it run for 5 minutes to set the motor brushes.  Turn the
motor off. Insure manometers are set to zero. Turn the high volume motor on.
      11.2.1.8 Record the time, in minutes, required to pass a known volume of air (approximately 200 to 300 ft3
of air for each  resistance plate) through the rootsmeter by using the rootsmeter's digital volume dial  and a
stopwatch.
      11.2.1.9  Record both manometer readings-orifice water manometer (AH) and rootsmeter mercury
manometer (A?) on Orifice Calibration Data Sheet (see Figure 7).

[Note: &His the sum of the difference from zero (0) of the two column heights.]

      11.2.1.10 Turn off the high volume motor.
      11.2.1.11  Replace the 5-hole resistance plate with the 7-hole  resistance plate.
      11.2.1.12 Repeat Sections 11.2.1.3 through  11.2.1.11.
      11.2.1.13  Repeat for each resistance plate.  Note results on Orifice Calibration Data Sheet (see Figure 7).
Only a minute is needed for warm-up of the motor. Be sure to tighten the orifice enough to eliminate any  leaks.
Also check the gaskets for cracks.

[Note: The placement of the orifice prior to the rootsmeter causes the pressure at the inlet of the rootsmeter
to be reduced below atmospheric conditions, thus causing the measured volume to  be incorrect.  The volume
measured by the rootsmeter must be corrected.]

      11.2.1.14 Correct the measured volumes on the Orifice Calibration Data Sheet:
                                               P -  A?   T.,
                                    v   =  V  (—-	v  stdi
                                    Vstd    Vm V        A    )
                                                   std      a

where:
        Vstd =  standard volume, std m3
         Vm =  actual volume measured by the rootsmeter, m3
          Pa =  barometric  pressure during calibration, mm Hg
         A? =  differential pressure at inlet to volume meter, mm Hg
         Pstd =  760 mm Hg
         Tstd=  273+25°C = 298K
          Ta =  ambient temperature during calibration, K.

      11.2.1.15 Record standard volume on Orifice Calibration Data Sheet.
      11.2.1.16  The standard flow rate as measured by the rootsmeter  can now be calculated using the following
formula:


                                           Qstd  =    std

where:
       Qstd = standard volumetric flow rate, std mVmin
         6 = elapsed time, min
Page 4A-12          Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

      11.2.1.17 Record the standard flow rates to the nearest 0.01 std mVmin.
      11.2.1.18 Calculate and record JAH (P1/Pstd)(298/T1) value for each standard flow rate.
      11.2.1. 19Plot each */AH  (P1/Pstd)(298/T1) value (y-axis) versus its associated standard flow rate (x-
axis) on arithmetic graph paper and draw a line of best fit between the individual plotted points.

[Note: This graph will be used in the field to determine standard flow rate.]

   11.2.2  Calibration of the High Volume Sampling System Utilizing Calibrated Orifice  Transfer
Standard
   For this calibration procedure, the following conditions are assumed in the field:

   •  The sampler is equipped with a valve to control sample flow rate.
   •  The sample flow  rate is determined by measuring the orifice pressure differential, using a Magnehelic
      gauge.
   •  The sampler is designed to operate at a standardized volumetric flow rate of 8 ftVmin (0.225 mVmin), with
      an acceptable flow rate range within 10 percent of this value.
   •  The transfer standard for the flow rate calibration is an orifice device. The flow rate through the orifice
      is determined by the pressure  drop caused by the orifice and is measured using a "U" tube water
      manometer or equivalent.
   •  The sampler and the orifice transfer standard are calibrated to standard volumetric flow rate units (scfm
      or scmm).
   •  An orifice transfer standard with calibration traceable to NIST is used.
   •  A "U" tube water manometer or equivalent, with a 0- to 16-inch range and a maximum scale division of
      0. 1 inch, will be used to measure the pressure in the orifice transfer standard.
   •  A Magnehelic gauge or equivalent, with a 9- to 100-inch range and a minimum scale division of 2 inches
      for measurements of the differential pressure across the sampler's orifice is used.
   •  A thermometer capable of measuring temperature over the range of 32° to 122 °F (0° to 50 °C) to±2°F
      (±1 °C) and referenced annually to a calibrated mercury thermometer is used.
   •  A portable aneroid barometer (or equivalent) capable of measuring ambient barometric pressure between
      500 and 800 mm Hg (19.5 and 31.5 in. Hg) to  the nearest mm Hg and referenced annually to a barometer
      of known accuracy is used.
   •  Miscellaneous handtools, calibration data sheets or station log book, and wide duct tape are available.

      1 1.2.2.1 Set up the calibration system as illustrated in Figure 8. Monitor the airflow through the sampling
system with a venturi/Magnehelic assembly, as illustrated in Figure 8. Audit the field sampling system once per
quarter using a flow rate transfer standard, as described in the EPA High  Volume-Sampling Method, 40 CVR 50,
Appendix B.  Perform a  single-point calibration before and after each  sample collection, using the procedures
described in Section 11.2.3.
      11.2.2.2 Prior to initial multi-point calibration, place an empty glass cartridge in the sampling head and
activate the sampling motor. Fully open the flow control valve and adjust the voltage variator so that a sample
flow rate corresponding to 1 10 percent of the desired flow rate (typically  0.20 to 0.28 mVmin) is indicated on the
Magnehelic gauge (based on the previously obtained multipoint calibration curve). Allow the motor to warm up
for 10 minutes and then adjust the flow  control valve to achieve the desire flow rate.  Turn off the sampler.
Record the ambient temperature and barometric pressure on the Field Calibration Data Sheet (see Figure 9).
      11.2.2.3 Place the orifice transfer standard on the sampling head and attach a manometer to the tap on
the transfer standard, as illustrated in Figure 8.  Properly align the retaining rings with the filter holder and secure
January 1999         Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-13

-------
Method TO-4A	Pesticides/PCBs

by tightening the three screw clamps.  Connect the orifice transfer standard by way of the pressure tap to a
manometer using a length of tubing.  Set the zero level of the manometer or Magnehelic. Attach the Magnehelic
gauge to the sampler venturi quick release connections. Adjust the zero (if needed) using the zero adjust screw
on face of the gauge.
      11.2.2.4 To leak test, block the orifice with a rubber stopper, wide duct tape, or other suitable means. Seal
the pressure port with a rubber cap or similar device.  Turn on the sampler.
Caution: Avoid running the sampler for too long a time with the orifice blocked.  This precaution will reduce
the chance that the motor will be overheated due to the lack of cooling air. Such overheating can shorten the
life of the  motor.
      11.2.2.5  Gently rock the orifice transfer standard and listen for a whistling sound that would indicate a
leak in the  system.  A leak-free system will not produce an upscale response on the sampler's Magnehelic.  Leaks
are usually caused either by damaged or missing gaskets by cross-threading and/or not screwing sample cartridge
together tightly. All leaks must be eliminated before proceeding with the calibration.  When the sample is
determined to be leak-free, turn off the sampler and unblock the orifice. Now remove the rubber stopper or plug
from the calibrator orifice.
      11.2.2.6 Turn the flow control valve to the fully open position and turn the sampler on. Adjust the flow
control valve until a Magnehelic reading of approximately  70 in.  is obtained.  Allow the Magnehelic and
manometer readings to stabilize and record these values on the orifice transfer Field Calibration  Data Sheet (see
Figure 9).
      11.2.2.7 Record the manometer reading under Yl and the Magnehelic reading under Y2 on the Field
Calibration Data Sheet. For the first reading, the Magnehelic should still be at 70 inches as set above.
      11.2.2.8  Set  the Magnehelic to 60 inches by using the sampler's flow control valve.  Record the
manometer (Yl) and Magnehelic (Y2) readings on the Field Calibration Data Sheet (see Figure 9).
      11.2.2.9  Repeat the above steps using Magnehelic settings of 50, 40, 30, 20, and 10 inches.
      11.2.2.10Turn the voltage variator to maximum power, open the flow control valve, and confirm that the
Magnehelic reads at least 100 inches. Turn off the sampler and confirm that the Magnehelic reads zero.
      11.2.2.11 Read and record the following parameters on the Field Calibration Data Sheet. Record the
following  on the calibration data sheet:
   Data, job number, and operator's signature;

   •   Sampler serial number;
   •   Ambient barometric pressure; and
   •   Ambient temperature.

      11.2.2.12 Remove the "dummy" cartridge and replace with a sample cartridge.
      11.2.2.13 Obtain the Manufacturer High Volume Orifice Calibration Certificate.
      11.2.2.14 If not performed by the manufacturer, calculate  values for each calibrator orifice static pressure
(Column 6, inches of water) on the manufacturer's calibration certificate using the following equation:
                                 ^/AH(Pa/760)(298/[Ta +  273])

where:
        Pa = the barometric pressure (mm Hg) at time of manufacturer calibration, mm Hg
        Ta = temperature at time of calibration, °C

      11.2.2.15  Perform a linear regression analysis using the values in Column 7 of the manufacturer High
Volume Orifice Calibration Certificate for flow rate (Qstd) as the "X" values and the calculated values as the Y
Page 4A-14           Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

values. From this relationship, determine the correlation (CC1), intercept (Bl), and slope (Ml) for the Orifice
Transfer Standard.
      11.2.2.16 Record these values on the Field Calibration Data Sheet (see Figure 9).
      11.2.2.17 Using the Field Calibration Data Sheet values (see Figure 9), calculate the Orifice Manometer
Calculated Values (Y3) for each orifice manometer reading using the following equation:

                                          Y3 Calculation

                                Y3 = [Yl(Pa/760)(298/{Ta + 273})f

      11.2.2.18 Record the values obtained in Column Y3 on the Field Calibration Data Sheet (see Figure 9).
      11.2.2.19 Calculate the Sampler Magnehelic Calculate Values (Y4) using the following equation:

                                          Y4 Calculation

                                Y4 = [Y2(Pa/760)(298/{Ta + 273 })f

      11.2.2.20 Record the value obtained in Column Y4 on the Field Calibration Data Sheet (see Figure 9).

      11.2.2.21 Calculate the Orifice Flow Rate (XI) in scm, using the following equation:

                                          XI Calculation

                                                Y3   -  Bl
                                        XI  =
                                                   Ml
      11.2.2.22 Record the values obtained in Column XI, on the Field Calibration Data Sheet (see Figure 9).
      11.2.2.23 Perform a linear regression of the values in Column XI (as X) and the values in Column Y4 (as
Y). Record the relationship for correlation (CC2), intercept (B2), and slope (M2) on the Field Calibration Data
Sheet.
      11.2.2.24 Using the following equation, calculate a set point (SP) for the manometer to represent a desired
flow rate:

          Set point (SP) = [(Expected Pa)/(Expected Ta)(Tstd/Pstd)] [M2 (Desired flow rate) + B2]2

where:

       Pa = Expected atmospheric pressure (Pa), mm Hg
       Ta = Expected atmospheric temperature (Ta), °C
      M2 = Slope of developed relationship
      B2 = Intercept of developed relationship
      Tstd = Temperature standard, 25 °C
      Pstd = Pressure standard, 760 mm Hg

      11.2.2.25  During monitoring, calculate a flow rate from the observed Magnehelic reading using the
following equations:
January 1999        Compendium of Methods for Toxic Organic Air Pollutants          Page 4A-15

-------
Method TO-4A	Pesticides/PCBs

                      Y5 = [Average Magnehelic Reading (AH) (Pa/Ta)(Tstd/Pstd)f


                                        X2  =  Y5  - B2
                                                   M2

where:

      Y5 = Corrected Magnehelic reading
      X2 = Instant calculated flow rate, scm

       11.2.2.26 The relationship in calibration of a sampling system between Orifice Transfer Standard and
flow rate through the sampler is illustrated in Figure 10.
    11.2.3 Single-Point Audit of the High Volume Sampling System Utilizing Calibrated Orifice Transfer
Standard
    Single point calibration checks are required as follows:

    •  Prior to the start of each 24-hour test period.
    •  After each 24-hour test period. The  post-test calibration check may serve as the pre-test calibration check
      for the next sampling period if the  sampler is not moved.
    •  Prior to sampling after a sample is moved.

For samplers, perform a calibration check for the operational flow rate before each 24-hour sampling event and
when required as outlined in the user quality assurance program. The purpose  of this check is to track the
sampler's calibration stability.  Maintain a control chart presenting the percentage difference between a sampler's
indicated and measured flow rates. This chart provides a quick reference of sampler flow-rate drift problems and
is useful for tracking the performance  of the sampler. Either the sampler log book or a data sheet will be used
to document flowcheck information.  This information includes, but is not  limited to, sampler and orifice transfer
standard serial number, ambient temperature, pressure conditions, and collected flow-check data.

In this subsection, the following is  assumed:

    •  The flow rate through a sampler is indicated by the orifice differential pressure;
    •  Samplers are designed to operate at an actual flow rate of 8 scfm, with a maximum acceptable flow-rate
      fluctuation range of ±10 percent of this value;
    •  The transfer standard will be an orifice device equipped with a pressure tap.  The pressure is measured
      using a manometer; and
    •  The orifice transfer standard's calibration relationship is in terms of standard volumetric flow rate (Qstd).

      11.2.3.1  Perform a single point flow audit check before and after each sampling period utilizing the
Calibrated Orifice Transfer Standard (see Section 11.2.1).
      11.2.3.2  Prior to single point audit, place a "dummy" glass cartridge in the sampling head and activate the
sampling motor. Fully open the flow control valve and adjust the voltage variator so that a sample flow rate
corresponding to 110 percent of the desired flow rate (typically 0.19 to  0.28  mVmin) is indicated on the
Magnehelic gauge (based on the previously obtained multipoint calibration curve). Allow the motor to warm up
for 10 minutes and then adjust the flow  control valve to achieve the desired flow rate.  Turn off the sampler.
Record the ambient temperature and barometric pressure on the Field Test Data Sheet (see Figure  11).
      11.2.3.3 Place the flow rate transfer standard on the sampling head.
Page 4A-16           Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

      11.2.3.4  Properly align the retaining rings with the filter holder and secure by tightening the 3 screw
clamps. Connect the flow rate transfer standard to the manometer using a length of tubing.
      11.2.3.5 Using tubing, attach 1 manometer connector to the pressure tap of the transfer standard. Leave
the other connector open to the atmosphere.
      11.2.3.6 Adjust the manometer midpoint by sliding the movable scale until the zero point corresponds with
the water meniscus. Gently shake or tap to remove any air bubbles and/or liquid remaining on tubing connectors.
(If additional liquid is required, remove tubing connector and add clean water.)
      11.2.3.7 Turn on high-volume motor and let run for 5 minutes.
      11.2.3.8 Record the pressure differential indicated, AH, in inches of water, on the Field Test Data Sheet.
Be sure stable AH has been established.
      11.2.3.9 Record the observed Magnahelic gauge reading, in inches of water, on the Field Test Data Sheet.
Be sure stable AM has been established.
      11.2.3.10    Using  previous  established  Orifice Transfer  Standard curve,  calculate   Qxs  (see
Section 11.2.2.23).
      11.2.3.11 This flow should be within ±10 percent of the sampler set point, normally, 8 ft3.  If not, perform
a new multipoint calibration of the sampler.
      11.2.3.12 Remove flow rate transfer standard and dummy adsorbent cartridge.

11.3 Sample Collection

   11.3.1 General Requirements
      11.3.1.1 The sampler should be located in an unobstructed area, at least 2 meters from any obstacle to air
flow. The exhaust hose should be stretched out in the downwind direction to prevent recycling of air into the
sample head.
      11.3.1.2  All cleaning and sample module loading and  unloading should be conducted in a controlled
environment, to minimize any chance of potential contamination.
      11.3.1.3 When new or when using the sampler at a different location, all sample contact areas need to be
cleared. Use triple rinses of reagent grade hexane contained in Teflon® rinse bottles.  Allow the solvent to
evaporate before loading the PUF modules.
   11.3.2 Preparing Cartridge for Sampling
      11.3.2.1 Detach the lower chamber of the cleaned sample head. While wearing disposable, clean, lint-free
nylon, or powder-free surgical gloves, remove  a clean glass adsorbent module from its  shipping container.
Remove the Teflon® end caps. Replace the end caps in the sample container to be reused after the sample has
been collected.
      11.3.2.2  Insert the glass module into the  lower chamber and tightly reattach the lower chambers to the
module.
      11.3.2.3 Using clean rinsed (with hexane) Teflon-tipped forceps, carefully place a clean conditioned fiber
filter atop the filter holder and secure in place by clamping the filter holder ring over the filter.   Place the
aluminum protective cover on top of the  cartridge head.  Tighten the 3 screw clamps. Ensure that all module
connections are tightly assembled. Place a small piece of aluminum foil on the ball-joint of the sample cartridge
to protect from back-diffusion of semi-volatile into the cartridge during transporting to the site.

[Note: Failure to do so could result in air flow leaks at poorly sealed locations which could affect sample
representativeness.]
January 1999         Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-17

-------
Method TO-4A	Pesticides/PCBs

      11.3.2.4 Place in a carrying bag to take to the sampler.
    11.3.3 Collection
      11.3.3.1 After the sampling system has been assembled, perform a single point flow check as described
in Sections 11.2.3.
      11.3.3.2 With the empty sample module removed from the sampler, rinse all sample contact areas using
reagent grade hexane in a Teflon® squeeze bottle. Allow the hexane to evaporate from the module before loading
the  samples.
      11.3.3.3 With the sample cartridge removed from the sampler and the flow control valve fully open, turn
the  pump on and allow it to warm-up for approximately 5 minutes.
      11.3.3.4 Attach a "dummy" sampling cartridge loaded with the exact same type of filter and PUF media
to be used for sample collection.
      11.3.3.5 Turn the sampler on and adjust the flow control valve to the desired flow as indicated by the
Magnehelic gauge reading determined in Section 11.2.2.24.  Once the flow is properly adjusted, take extreme care
not to inadvertently alter its setting.
      11.3.3.6 Turn the sampler off and remove the "dummy" module.  The sampler is now ready for field use.
      11.3.3.7 Check the zero reading of the sampler Magnehelic.  Record the ambient temperature, barometric
pressure, elapsed time meter setting, sampler serial number, filter number, and PUF cartridge number on the Field
Test Data  Sheet (see Figure 11). Attach the loaded sampler cartridge to the sampler.
      11.3.3.8 Place the voltage variator and flow control valve at the settings used in Section 11.3.2, and the
power switch. Activate the elapsed time meter and record the start time.  Adjust the flow (Magnehelic setting),
if necessary, using the flow control valve.
      11.3.3.9 Record the Magnehelic reading every  6 hours during the sampling period.  Use the calibration
factors (see Section 11.2.2.24) to calculate the desired flow rate.  Record the ambient temperature, barometric
pressure, and Magnehelic reading at the beginning and during sampling period.
    11.3.4 Sample Recovery
      11.3.4.1 At the end of the desired sampling period, turn the power off.  Carefully remove the sampling
head containing the filter and adsorbent cartridge.  Place the protective "plate" over the filter to protect cartridge
during transport to  clean recovery area.  Also, place a piece of aluminum foil around the bottom of adsorbent
sampler head.
      11.3.4.2 Perform a final calculated sampler flow check using the calibration orifice, as described in Section
11.3.2. If calibration deviates by more than 10 percent from initial reading, mark the flow data for that sample
as suspect and inspect and/or remove from service, record results on Field Test Data Sheet, Figure 11.
      11.3.4.3 Transport adsorbent sampler head to a clean recovery area.
      11.3.4.4  While wearing disposable  lint free nylon or powder-free surgical gloves, remove the PUF
cartridge from the  lower module chamber and lay it on the  retained aluminum foil in which the sample was
originally wrapped.
      11.3.4.5  Carefully remove the glass fiber filter  from the upper chamber using clean Teflon®-tipped
forceps.
      11.3.4.6 Fold the filter in half twice (sample side inward) and place it in the glass cartridge atop the PUF.
      11.3.4.7 Wrap the combined samples in the original hexane rinsed aluminum foil, attached Teflon® end
caps and place them in their original aluminum sample container. Complete a sample label and affix it to the
aluminum shipping container.
      11.3.4.8 Chain-of-custody should be maintained for all samples. Store the containers under dry ice and
protect from UV light to prevent possibly photo-decomposition of collected analytes. If the time span between
sample collection and laboratory analysis is to exceed 24 hours, refrigerate sample at 4°C.
      11.3.4.9 Return at least 1 field filter/PUF blank to the laboratory  with each group of samples. Treat a field
blank exactly as the sample except that no air is drawn through the filter/adsorbent cartridge assembly.
Page 4A-18           Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

      11.3.4.10 Ship and store field samples chilled (<4°) (blue ice is acceptable) until receipt at the analytical
laboratory, after which they should be refrigerated at less than or equal to 4°C. Extraction must be performed
within 7 days of sampling and analysis within 40 days of extraction.
12. Sample Extraction Procedure

[Note: Sample extraction should be performed under a properly ventilated hood.]

12.1 Sample Extraction

    12.1.1 All samples should be extracted within 1 week after collection. All samples should be stored at <4°C
until extracted.
    12.1.2 All glassware should be washed with a suitable detergent; rinsed with deionized water, acetone, and
hexane; rinsed again with deionized water; and fired in an oven (500°C).
    12.1.3  Prepare a spiking solution for determination of extraction efficiency.  The spiking solution should
contain one or more surrogate compounds that have chemical structures and properties similar to those of the
analytes of interest.  Octachloronaphthalene (OCN) and dibutylchlorendate have been used as surrogates for
determination of organochlorine pesticides by GC with an BCD.  Tetrachloro-m-xylene and decachlorobiphenyl
can also be used together to insure recovery of early and late eluting compounds. For organophosphate pesticides,
tributylphosphate or triphenylphosphate may be employed as  surrogates.  The surrogate solution should be
prepared so that addition of 100 /j,L into the PUF plug results in an extract containing the surrogate compound
at the high end of the instrument's calibration range.  As an example, the spiking solution for OCN is prepared
by dissolving 10 mg of OCN in  10 mL of 10% acetone in n-hexane, followed by serial dilution n-hexane to
achieve a final spiking solution of OCN is 1 //g/mL.

[Note: Use the recoveries of the surrogate compounds to monitor for unusual matrix effects and gross sample
processing errors.   Evaluate surrogate recovery for acceptance by determining whether the measured
concentration falls within the acceptance limits of 60-120 percent.]

    12.1.4 The extracting solution (10% diethyl ether/hexane) is prepared by mixing 1800 mL of freshly opened
hexane and 200 mL of freshly opened diethyl ether (preserved with ethanol) to a flask.
    12.1.5 All clean glassware, forceps, and other equipment to be used should be rinsed with 10% diethyl ether/
hexane and placed on rinsed (10% diethyl ether/hexane) aluminum foil until use.  The condensing towers should
also be rinsed with 10% diethyl ether/hexane.  Then add 700 mL of 10% diethyl ether/hexane to the 1,000 mL
round bottom flask and add up to three boiling granules.
    12.1.6  Using precleaned (i.e., 10% diethyl ether/hexane Soxhlet extracted) cotton gloves, the filter/PUF
cartridge is removed from the sealed container, the PUF removed from the glass cartridge, and the filter/PUF
together  are placed into the 300 mL Soxhlet extractor using prerinsed forceps.
    12.1.7  Before extraction begins, add 100 uL of the OCN solution directly to the top of the PUF plug.

[Note: Incorporating a known concentration of the solution onto the sample provides a quality assurance
check to determine recovery efficiency of the extraction and analytical processes.]

    12.1.8 Connect the Soxhlet extractor to the 1,000 mL boiling flask and condenser.  Wet the glass joints with
10% diethyl ether/hexane to ensure a tight seal between the fittings. If necessary, the PUF plug can be adjusted
January 1999         Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-19

-------
Method TO-4A	Pesticides/PCBs

using forceps to wedge it midway along the length of the siphon.  The above procedure should be followed for
all samples, with the inclusion of a blank control sample.

    12.1.9 The water flow to the condenser towers of the Soxhlet extraction assembly should be checked and the
heating unit turned on. As the samples boil, the Soxhlet extractors should be inspected to ensure that they are
filling and siphoning properly (4 to 6 cycles/hour). Samples should cycle for a minimum of 16 hours.
    12.1.10 At the end of the extracting process (minimum of 16 hours), the heating unit is turned off and the
sample cooled to room temperature.
    12.1.11 The extracts are then concentrated to 5 mL using a Kuderna-Danish (K-D) apparatus.  The K-D is
set up, assembled with concentrator tubes, and rinsed.  The lower end of the filter tube is packed with glass wool
and filled with sodium sulfate to a depth of 40 mm.  The filter tube is then placed in the neck of the K-D.  The
Soxhlet extractors and boiling flasks are carefully removed from the condenser towers and the remaining solvent
is drained into each boiling flask.  Sample extract is carefully poured through the filter tube into the K-D. Each
boiling flask is rinsed three times by swirling hexane along the sides. Once the sample has drained, the filter tube
is rinsed down with hexane. Each Snyder column is attached to the K-D and rinsed to wet the joint for a tight
seal. The complete K-D apparatus is placed on a steam bath and the sample is evaporated to approximately 5
mL.

[Note: Do not allow samples to evaporate to dryness.]

Remove sample from the steam bath, rinse the Snyder column with a minimum of hexane, and allow to cool.
Adjust sample volume  to 10 mL  in a concentrator tube, close with a glass stopper, and seal with TFE
fluorocarbon tape. Alternatively, the sample may be quantitatively transferred (with concentrator tube rinsing)
to prescored vials  and brought up to final volume. Concentrated extracts are stored at <4°C until analyzed.
Analysis should occur no later than 40 days after sample extraction.

12.2 Sample Cleanup

    12.2.1  If only polar compounds are sought, an alumina cleanup procedure is appropriate. Before cleanup,
the  sample extract is carefully reduced to 1 mL using a gentle stream of clean nitrogen.
    12.2.2 A glass chromatographic column (2-mm ID. x 15-cm long) is packed with alumina (7), activity grade
IV,  and rinsed with approximately 20 mL of n-hexane. The concentrated sample extract is placed on the column
and eluted with 10 mL of n-hexane at a rate of 0.5 mL/minute.  The eluate volume is adjusted to exactly 10 mL
and analyzed as per Section 13.
    12.2.3 If both PCBs and common pesticides are sought, alternate cleanup procedures (8,9) may be required
(i.e., silicic acid).
    12.2.4 Finally, class separation and improved specificity can be achieved by column clean-up and separation
on Florisil (9).
Page 4A-20          Compendium of Methods for Toxic Organic Air Pollutants         January 1999

-------
Pesticides/PCBs                                                                    Method TO-4A
13. Analytical Procedure

13.1 Analysis of Organochlorine Pesticides by Capillary Gas Chromatography with Electron Capture
Detector (GC/ECD)

[Note:  Organochlorine pesticides, PCBs and many nonchlorinatedpesticides are responsive to electron
capture detection (see Table 1). Most of these compounds can be analyzed at concentration of I to 50 ng/mL
by GC/ECD.  The following procedure is appropriate. Sampling and analytical methods that have been used
to determine pesticides and PCBs collected from air using a modification of this methodology have been
published (14-22).]

    13.1.1 Select GC column (e.g., 0.3-mm by 30-m DB-5 column) and appropriate GC conditions to separate
the target analytes. Typical operating parameters for this column with splitless injection are:  Carrier gas-
chromatography grade helium at a flow rate of 1 to 2 mL/min and a column head pressure of 7 to 9 psi (48 to
60 kPa); injector temperature of 250°C; detector temperature of 350°C; initial oven temperature of 50°C held
for 2.Omin., ramped at 15°C/minto 150°Cfor 8min, ramped at 10°C/min to 295 °C then held for 5 min; purge
time of 1.0 min. A typical injection volume is 2 to 3 ,uL.
    13.1.2  Remove sample extract from refrigerator and allow to warm to room temperature.
    13.1.3 Prepare standard solution from reference materials of known purity.  Analytically pure standards of
Organochlorine pesticides and PCBs are available from several commercial sources.
    13.1.4 Use the standard solutions of the various compounds of interest to determine relative  retention times
(RRTs) to an internal standard such as p,p'-DDE, aldrin or octachloronaphthalene. Use 1 to 3-//L injections or
other appropriate volumes.
    13.1.5 Determine detector linearity by injecting standard solutions of three different concentrations (amounts)
that bracket the range of analyses.  The  calibration is considered linear if the relative standard deviation (RSD)
of the three response factors for the three standards is 20 percent or less.
    13.1.6 Calibrate the system with a minimum of three levels of calibration standards in the linear range. The
low  standard should be near the analytical method detection limit.  The calibration is considered linear if the
relative standard deviation (RSD) of the three response factors for the three standards is 20 percent or less. The
initial calibration should be verified by  the  analysis of a standard from an independent source.  Recovery of 85
to 115 percent is acceptable.  The initial calibration curve should be verified at the begining of each day and after
every ten samples by the analysis of the  midpoint standard; an RPD of 15% or less is acceptable for continuing
use of the initial calibration curve.
    13.1.7  Inject  1 to 3 /jL of sample extract. Record volume injected to the nearest 0.05 /jL.
    13.1.8  A typical BCD response for a mixture of single component pesticides using a capillary column is
illustrated in Figure 12. If the response (peak height or area) exceeds the calibration range, dilute the extract and
reanalyze.
    13.1.9 Quantify PCB mixtures by comparison of the total heights or areas of GC peaks (minimum of five)
with the corresponding peaks in the best-matching standard. Use Aroclor 1242 for early-eluting PCBs and either
Aroclor 1254 or Aroclor 1260 as appropriate for late-eluting PCBs.
    13.1.10   If both PCBs and  Organochlorine pesticides are  present in the same  sample,  use column
chromatographic separation on silicic acid (8,9) prior to GC analysis.
    13.1.11 If polar compounds are present that interfere with GC/ECD analysis, use column chromatographic
cleanup or alumina (7), activity grade IV, in accordance with Section 12.2.
January 1999         Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-21

-------
Method TO-4A	Pesticides/PCBs

    13.1.12 For confirmation use a second GC column such as DB-608. All GC procedures except GC/MS
require second column confirmation.
    13.1.13 For improved resolution use a capillary column such as an 0.25-mm I.D. x 30-m DB-5 with 0.25 ,um
film thickness. The following conditions are appropriate.

    •   Helium carrier gas at 1 mL/min.
    •   Column temperature program, 90°C (4 min)/16°C/min to  154°C/4°C/min to 270°C.
    •   Detector, 63Ni BCD at 350°C.
       Make up gas, nitrogen, or 5% methane/95% argon at 60 mL/min.
    •   Splitless injection, 2 ,uL maximum.
       Injector temperature, 220°C.

    13.1.14 Class separation and improved specificity can be achieved by column chromatographic separation
on Florisil (9).
    13.1.15 A Hall electrolytic conductivity detector (HECD) operated in the reductive mode may be substituted
for the BCD for improved specificity.  Sensitivity, however, will be reduced by at least an order of magnitude.

13.2   Analysis of Organophosphorus Pesticides by Capillary Gas Chromatography with  Flame
Photometric or Nitrogen-Phosphorus Detectors (GC/FPD/NPD)

[Note: Organophosphorus pesticides are responsive to flame photometric and nitrogen-phosphorus (alkali
flame ionization)  detection. Most of these compounds can  be analyzed at concentrations of 50 to 500 ng/mL
using either of these detectors.]

    13.2.1 Procedures given in Section 13.1.1 through 13.1.9 and Section 13.1.13 through 13.1.14 apply, except
for the selection of surrogates.
    13.2.2  Use tributylphosphate, triphenylphosphate, or other suitable compound(s) as surrogates to verify
extraction efficiency and to determine RRTs.

13.3  Analysis of Carbamate and Urea Pesticides by Capillary Gas Chromatography with Nitrogen-
Phosphorus Detector

    13.3.1  Trazine, carbamate, and urea pesticides may be determined by capillary GC (DB-5, DB-17, or
DB-1701 stationary phase) using nitrogen-phosphorus detection or MS-SIM with  detection limits in the 0.05 to
0.2 //L/mL range. Procedures given in Section 13.1.1 through 13.1.9 and Section 13.1.13 through 13.1.14 apply,
except for the selection of surrogates, detector, and make up gas.
    13.3.2 Thermal degradation may be minimized by reducing the  injector temperature to 200°C.  HPLC may
also be used, but detection limits will be higher (1 to 5 //g/mL).
    13.3.3    N-methyl carbamates  may be determined  using reverse-phase  high performance liquid
Chromatography  (HPLC) (C-18)  (Section 13.4) and post-column derivization with o-phthaldehyde  and
fluorescence detection (EPA Method 531). Detection limits of 0.01 to 0.1 ,ug/mL can be achieved.
Page 4A-22          Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs	Method TO-4A

13.4  Analysis of Carbamate, Urea, Pyrethroid, and Phenolic Pesticides by High Performance Liquid
Chromatography (HPLC)

[Note: Many carbamate pesticides, urea pesticides, pyrethrins, phenols, and other polar pesticides may be
analyzed by high HPLC with fixed or variable wavelength UV detection. Either reversed-phase or normal
phase chromatography may be used.  Detection limits are 0.2 to 10 i^g/mL of extract.]

    13.4.1 Select HPLC column (i.e., Zorbax-SIL, 46-mm I.D. x 25-cm, or ^-Bondapak CIS, 3.9-mm x 30-cm,
or equivalent).
    13.4.2 Select solvent system (i.e., mixtures of methanol or acetonitrile with water or mixtures of heptane or
hexane with isopropanol).
    13.4.3 Follow analytical procedures given in Sections 13.1.2 through 13.1.9.
    13.4.4   If interferences are present, adjust  the  HPLC  solvent system  composition or use  column
chromatographic clean-up with silica gel, alumina,  or Florisil (9).
    13.4.5  An electrochemical detector may be used to improve sensitivity for some ureas, carbonates, and
phenolics.  Much more care is required in using this detector, particularly in removing dissolved oxygen from the
mobile phase and sample extracts.
    13.4.6 Chlorophenol (di- through penta-) may be analyzed by GC/ECD or GC/MS after derivatization with
pentafluorobenzylbromide (EPA Method 604).
    13.4.7 Chlorinated phenoxyacetic acid herbicides and pentachlorophenol can be analyzed by GC/ECD or
GC/MS after derivatization with diazomethane (EPA Method  515). DB-5 and DBJ-1701 columns  (0.25-mm
I.D. x 30-m) at 60 to 300°C/4°C per min have been found to perform well.

13.5  Analysis of Pesticides and PCBs  by Gas Chromatography with  Mass Spectrometry Detection
(GC/MS)

[Note: A mass spectrometer operating in the selected ion monitoring mode is useful for confirmation and
identification of pesticides.]

    13.5.1  A mass  spectrometer operating in select ion monitoring (SIM) mode can be used as a sensitive
detector for multi-residue determination of a wide variety of pesticides.  Mass spectrometers are now available
that provide detection limits comparable to nitrogen-phosphorus and electron capture detectors.
    13.5.2 Most of the pesticides shown in Table 1 have been successfully determined by GC/MS-SIM.  Typical
GC operating parameters are as described in  Section 13.1.1.
    13.5.3 The mass spectrometer is typically operated using  positive ion electron impact ionization (70 eV).
Other instrumental parameters are instrument specific.
    13.5.4 p-Terphenyl-d14 is commonly used as a surrogate for GC/MS analysis.
    13.5.5  Quantification is typically performed using an internal standard method.  1,4-Dichlorobenzene,
naphthalene-d8, acenaphthene-d10, phenanthrene-d10, chrysene-d12  and perylene-d12 are commonly used as internal
standards.  Procedures given in Section 13.1.1 through 13.1.9 and Section 13.1.13 through 13.1.14 apply, except
for the selection of surrogates, detector, and make up gas.
    13.5.6 See ASTM Practice D 3687 for injection technique, determination of relative retention times, and
other procedures pertinent to GC and HPLC analyses.
January 1999         Compendium of Methods for Toxic Organic Air Pollutants          Page 4A-23

-------
Method TO-4A	Pesticides/PCBs

13.6 Sample Concentration

    13.6.1  If concentrations are too low to detect by the analytical procedure of choice, the extract may be
concentrated to  1 mL or 0.5 mL by carefully controlled evaporation under an inert atmosphere.  The following
procedure is appropriate.
    13.6.2  Place K-D concentrator tube in a water bath and analytical evaporator (nitrogen blow-down)
apparatus. The  water bath temperature should be from 25 °C to 50 °C.
    13.6.3 Adjust nitrogen flow through hypodermic needle to provide a gentle stream.
    13.6.4 Carefully lower hypodermic needle into the concentrator tube to a distance of about 1 cm above the
liquid level.
    13.6.5 Continue to adjust needle placement as liquid level decreases.
    13.6.6 Reduce volume to slightly below desired level.
    13.6.7 Adjust to final volume by carefully rinsing needle tip and concentrator tube well with solvent (usually
n-hexane).
14. Calculations

14.1 Determination of Concentration

    14.1.1 The concentration of the analyte in the extract solution can be taken from a standard curve where peak
height or area is plotted linearly against concentration in nanograms per milliliter (ng/mL).  If the detector
response is known to be linear, a single point is used as a calculation constant.
    14.1.2 From the standard curve, determine the nanograms of analyte standard equivalent to the peak height
or area for a particular compound.
    14.1.3 Ascertain whether the field blank is contaminated. Blank levels should not exceed 10 ng/sample for
organochlorine pesticides or 100 ng/sample for PCBs and other pesticides.  If the blank has been contaminated,
the sampling series must be held suspect.

14.2 Equations

    14.2.1 Quantity of the compound in the sample (A) is calculated using the following equation:

                                                  A x V
                                      A  =  1000
       where:
                  A =  total amount of analyte in the sample, ng.
                 As =  calculated amount of material injected onto the chromatograph based on calibration
                       curve for injected standards, ng.
                 Ve=  final volume of extract, mL.
                 V; =  volume of extract injected, ,uL.
               1000 =  factor for converting microliters to milliliters.
Page 4A-24           Compendium of Methods for Toxic Organic Air Pollutants         January 1999

-------
Pesticides/PCBs	Method TO-4A

   14.2.2 The extraction efficiency (EE) is determined from the recovery of surrogate spike as follows:

                                       EE(%) =  — [100]
                                                 O
     where:
            EE =  extraction efficiency, %
              S =  amount of spike recovered, ng.
             Sa =  amount of spike added to plug, ng.

The extraction efficiency (surrogate recovery) must fall between 60-120% to be acceptable.

   14.2.3 The total volume of air sampled under ambient conditions is determined using the following equation:

                                             £  (TiXFj)
                                              1000 L/m3

     where:
             Va =  total volume of air sampled, m3.
              T; =  length of sampling segment between flow checks, min.
              F; =  average flow during sampling segment, L/min.
   14.2.4 The air volume is corrected to EPA standard temperature (25 ° C) and standard pressure (760 mm Hg)
as follows:
                                V  =V
                                       a (, 760 mm Hg)  (  tA  j

     where:
              Vs =  volume of air at standard conditions (25 °C and 760 mm Hg), std. m3.
              Va =  total volume of air sampled, m3.
              Pb =  average ambient barometric pressure, mm Hg.
              Pw =  vapor pressure of water at calibration temperature, mm Hg.
              tA =  average ambient temperature, °C + 273.

   14.2.5 If the proper criteria for a sample have been met, concentration of the compound in a standard cubic
meter of air sampled is calculated as follows:

                                     Ca(ng/std.  m3) =  ^
January 1999        Compendium of Methods for Toxic Organic Air Pollutants          Page 4A-25

-------
Method TO-4A	Pesticides/PCBs

    If it is desired to convert the air concentration value to parts per trillion (ppt) in dry air at standard
temperature and pressure (STP), the following conversion is used:

                                          ppt=0.844(Ca)

    The air concentration can be converted to parts per trillion (v/v) in air at STP as follows:
                                               (24.45)  (Ca)"
                                       pptv =
                                                  (MW)

      where:
            MW =  molecular weight of the compound of interest, g/g-mole.

    14.2.6 If quantification is performed using an internal standard, a relative response factor (RRF) is calculated
by the equation:

                                         RRF  =

      where:
(U(cs)
               Is =  integrated area of the target analyte peak, counts.
               I1S =  integrated area of the internal standard peak, counts.
              C1S =  concentration of the internal standard, ng/juL.
              Cs =  concentration of the analyte, ng/uL.

    14.2.7  The concentration of the analyte (Ca) in the sample is then calculated as follows:
                                          c
                                                (RRF)(y

      where:
               Is =  integrated area of the target analyte peak, counts.
            RRF =  relative response factor (see Section 14.2.7).

15. Performance Criteria and Quality Assurance

[Note: This section summarizes required quality assurance (QA) measures and provides guidance concerning
performance criteria that should be achieved within each laboratory.]
Page 4A-26           Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs                                                                     Method TO-4A
15.1 Standard Operating Procedures (SOPs)

    15.1.1 Users should generate SOPs describing the following activities accomplished in their laboratory: (1)
assembly, calibration, and operation of the sampling system, with make and model of equipment used; (2)
preparation, purification, storage, and handling of sampling cartridges, (3) assembly, calibration, and operation
of the analytical system, with make and model of equipment used; and (4) all aspects of data recording and
processing, including lists of computer hardware and software used.
    15.1.2 SOPs should provide specific stepwise instructions and should be readily available to, and understood
by, the laboratory personnel conducting the work.

15.2 Process, Field, and Solvent Blanks

    15.2.1  One filter/PUF cartridge from each batch of approximately twenty should be analyzed, without
shipment to the field, for the compounds of interest to serve as a process blank.
    15.2.2 During each sampling episode, at least one filter/PUF cartridge should be shipped to the field and
returned, without drawing air through the sampler, to serve as a field blank.
    15.2.3  Before each sampling episode, one PUF plug from each batch of approximately twenty should be
spiked with a known amount of the standard solution. The spiked plug will remain in a  sealed container and will
not be used during the sampling period.  The spiked plug is extracted  and analyzed with the other samples. This
field spike acts as a quality assurance  check to  determine matrix spike recoveries and to indicate sample
degradation.
    15.2.4 During the analysis of each batch of samples, at least one solvent process blank (all steps conducted
but no filter/PUF cartridge included) should be carried through the procedure and analyzed.
    15.2.5 Levels for process, field and solvent blanks should not exceed 10 ng/sample for single components
or 100 ng/sample for multiple component mixtures (i.e., for organochlorine pesticides and PCBs).

15.3 Method Precision and Bias

    15.3.1 Precision and bias in this type of analytical procedure are dependent upon the precision and bias of
the analytical procedure for each compound of concern, and the precision and bias of the sampling process.
    15.3.2  Several  different parameters involved in both the sampling and analysis  steps  of this method
collectively determine the precision and bias with which each compound is detected.   As the volume of air
sampled is increased, the sensitivity of detection increases proportionately within limits set by: (a) the retention
efficiency for each  specific  component trapped  on the polyurethane foam plug,  and (b) the background
interference associated with the analysis of each specific component at a given site sampled.  The sensitivity of
detection of samples recovered by extraction depends on:  (a) the inherent response of the particular GC detector
used in the determinative step, and (b) the extent to which the sample is concentrated for analysis. It is the
responsibility of the analyst(s) performing the sampling and analysis steps to adjust parameters so that the
required detection limits can be obtained.
    15.3.3 The reproducibility of this method for most compounds for which it has  been evaluated has been
determined to range  from ±5 to ±30% (measured as  the relative  standard deviation)  when replicate sampling
cartridges are used (N>5).  Sample recoveries for individual compounds generally fall  within the range of 90 to
110%, but recoveries ranging from 65 to 125% are considered acceptable.
January 1999         Compendium of Methods for Toxic Organic Air Pollutants          Page 4A-27

-------
Method TO-4A	Pesticides/PCBs

15.4 Method Safety

    15.4.1 This procedure may involve hazardous materials, operations, and equipment.  This method does not
purport to address all of the safety problems associated with its use.
    15.4.2  It is the users responsibility to consult and establish appropriate safety and health practices and
determine the applicability of regulatory limitations prior to the implementation of this procedure. This should
be part of the users  SOP manual.
16. References

1. Riggin, R. M., Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient
Air, U. S. Environmental Protection Agency, Environmental Monitoring Systems Laboratory, Quality Assurance
Division, Research Triangle Park, NC, EPA-600/4-84-041, April 1984.

2. Winberry, W. T. Jr., et al., "Determination of Benzo(a)Pyrene and Other Polynuclear Aromatic Hydrocarbons
(PAHs) in Ambient Air Using Gas Chromatographic (GC) and High Performance Liquid Chromatographic
(HPLC) Analysis:  Method TO-13," in Compendium of Methods for the Determination of Toxic Organic
Compounds in Ambient Air,  Second Supplement, U.  S. Environmental Protection Agency, Environmental
Monitoring Systems Laboratory, Quality Assurance Division, Research Triangle Park, NC, EPA-600/4-89-018,
March 1989.

3. Winberry, W. T.  Jr., et al., "Determination of Organochlorine Pesticides in Indoor Air: Method IP-8," in
Compendium of Methods for the Determination of Air Pollutant in Indoor Air, U. S. Environmental Protection
Agency, Research Triangle Park, NC, EPA-600/4-90-010,  May 1990.

4. "Standard Practice for Sampling and Analysis of Pesticides and Polychlorinated Biphenyls in Air," Annual
Book of ASTMStandards, Method D4861-94, ASTM, Philadelphia, PA.

5. Lewis, R., and MacLeod, K., "Portable Sampler for Pesticides and Semi-Volatile Industrial Organic Chemicals
in Air," Anal. Chem.,Vol 54, 1982, pp. 310-315.

6. Winberry, W. T. Jr., et al., "Determination of Organochlorine Pesticides in Ambient Air Using Low Volume
Polyurethane Foam (PUF) Sampling with Gas Chromatography/Electron Capture Detector (GC/ECD):  Method
TO-10, in Compendium of Methods for the Determination of Toxic Organic Compounds  in Ambient Air,
Second Supplement, U. S. Environmental Protection Agency, Research Triangle Park, NC, EPA-600/4-89-018,
March 1989.

7. Lewis, R., and Brown, A.,  and Jackson, M., "Evaluation of Polyurethane Foam for Sampling of Pesticides,
Polychlorinated Biphenyls and Polychlorinated Napththalenes in Ambient Air," Anal. Chem., Vol. 49, 1977, pp.
1668-1672.

8. Armour, J., and Burke, J.,  "Method for Separating Polychlorinated Biphenyls from DDT and Its Analogs,"
Journal of the Association of Official Analytical Chemists, Vol.  53, No. 4, 1970, pp. 761-768.
Page 4A-28          Compendium of Methods for Toxic Organic Air Pollutants         January 1999

-------
Pesticides/PCBs	Method TO-4A

9. Manual of Analytical Methods for the Analysis of Pesticides in Human and Environmental Samples, U. S.
Environmental Protection Agency, Research Triangle Park, NC, EPA-600/8-80-038, June 1980 (NTIS No.
PB82-208752).

10.  Carcinogens - Working with Carcinogens, Department of Health, Education, and Welfare, Public Health
Service, Center for Disease Control, National Institute for Occupational Safety and Health, Publication No. 77-
206, August 1977.

11.   OSHA Safety and Health Standards, General Industry, (29CFR1910), Occupational Safety and Health
Administration, OSHA, 2206, Revised, January  1976.

12.   "Safety in Academic Chemistry Laboratories," American Chemical Society Publication, Committee on
Chemical Safety, 3rd Edition, 1979.

13.  Kogan, V., Kuhlman, M., Coutant, R, and Lewis, R., "Aerosol Filtration in Sorbent Beds," Journal of the
Air and Waste Management Association,Vol. 43, 1993, pp. 1367-1373.

14.   Lewis, R., and Lee, R., "Air Pollution from Pesticide Sources, Occurrences and Dispersion," in: Air
Pollution from Pesticides and Agricultural Processes, Lee, R., Editor, CRC Press, Boca Raton, FL, 1976,
pp.  51-94.

15.  Lewis, R, "Problem Associated with Sampling for Semi-Volatile Organic Chemicals in Air," in Proceedings
of the  1986 EPA/APCA Symposium on Measurement of Toxic  Air Pollutants,  Air and Waste Management
Association, Pittsburgh, PA, 1986, pp.  134-145.

16.   Camann, D., Harding, J., and Lewis, R., "Trapping of Particle-Associated Pesticides in Indoor Air by
Polyurethane Foam and  Evaporation of Soil Track-In  as a Pesticide Source,"  in:  Indoor Air '90, Vol. 2,
Walkinshaw, D., Editor, Canada Mortgage and Housing Corp., Ottawa,  1990, pp. 621-626.

17.  Marple, V., Rubow, K., Turner, W., and Spengler, J., "Low Flow Rate Sharp Cut Impactors for Indoor Air
Sampling Design and Calibration," Journal of the Air Pollution Control Association. Vol. 37, 1987, pp. 1303-
1307.

18.   Hsu,  J., Wheeler, H., Camann, D., Shatterberg,  H., Lewis, R., and Bond,  A., "Analytical Methods for
Detection of Non-Occupational Exposure to Pesticides," Journal ofChromatographic Science, Vol. 26, 1988,
pp.  181-189.

19.   Lewis, R. G., and Jackson, M. D., "Modification  and Evaluation of  a High-Volume Air Sampler for
Pesticides and Semi-Volatile Industrial Organic Chemicals," Anal. Chem., 54, 592-594, 1982.

20.   Lewis, R.  G., Jackson, M. D., and MacLeod, K. E., "Protocol for Assessment of Human Exposure to
Airborne Pesticides," U. S. Environmental Protection Agency, Research Triangle Park, NC, EPA-600/2-80-180,
May 1980.

21.  Riggin, R. M., Technical Assistance Document for Sampling and Analysis of Toxic Organic Compounds
in Ambient Air, U. S. Environmental Protection Agency, Research  Triangle Park, NC, EPA-600/4-83-027, June
1983.
January 1999        Compendium of Methods for Toxic Organic Air Pollutants           Page 4A-29

-------
Method TO-4A	Pesticides/PCBs

22. Longbottom, J. E., and Lichtenberg, J. J., "Methods for Organic Chemical Analysis of Municipal and
Industrial Wastewater," U. S. Environmental Protection Agency, Cincinnati, OH, EPA-600/4-82-057, May 1982.
Page 4A-30          Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs
                                                               Method TO-4A
             TABLE 1.  COMPOUNDS FOR WHICH PROCEDURE HAS BEEN TESTED1
Compound
Alachlor
Aldrin
Allethrin
Aroclor 1242
Aroclor 1254
Aroclor 1260
Atrazine
Bendiocarb
BHC (a- and (3-Hexachlorocyclohexanes)
Captan
Carbaryl
Carbofuran
Chlordane, technical
Chlorothalonil
Chlorotoluron
Chlorpyritos
2,4-D esters and salts
Dacthal
p,p-'DDT
p,p-'DDE
Diazinon
Dicloran
Dieldrin
Dicofol
Dicrotophos
Diuron
Ethyl parathion
Fenvalerate
Fluometuron
Recommended
Analysis2
GC/ECD
GC/ECD
HPLC/UV
GC/ECD
GC/ECD
GC/ECD
GC/NPD
HPLC/UV
GC/ECD
GC/ECD
HPLC/UV
HPLC/UV
GC/ECD
GC/ECD
HPLC/UV
GC/ECD
GC/ECD
GC/ECD
GC/ECD
GC/ECD
GC/NPD or FPD
GC/ECD
GC/ECD
GC/ECD
HPLC/UV
HPLC/UV
GC/NPD or FPD
HPLC/UV
HPLC/UV
Compound
Folpet
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Lindane (y-BHC)
Linuron
Malathion
Methyl parathion
Methoxychlor
Metolachlor
Mexacarbate
Mirex
Monuron
Trans-nonachlor
Oxychlordane
Pentachlorobenzene
Pentachlophenol
Permethrin (cis and trans)
o-Phenylphenol
Phorate
Propazine
Propoxur (Baygon)
Pyrethrin
Resmethrin
Ronnel
Simazine
Terbuthiuron
Trifluralin

Recommended
Analysis
GC/ECD
GC/ECD
GC/ECD
GC/ECD
GC/ECD
HPLC/UV
GC/NPD or FPD
GC/NPD or FPD
GC/FCD
GC/ECD
GC/FCD
GC/ECD
HPLC/UV
GC/ECD
GC/ECD
GC/ECD
GC/ECD
HPLC/UV
HPLC/UV
GC/NPD or FPD
GC/NPD
HPLC/UV
HPLC/UV
HPLC/UV
GC/ECD
HPLC/UV
HPLC/UV
GC/ECD

1  The following recommendations are specific for that analyte for maximum sensitivity.
2  GC = gas chromatography; ECD = electron capture detector, FPD = flame photometric detector; HPLC = high performance
  liquid chromatography; NPD = nitrogen-phosphorus detector; UV = ultraviolet absorption detector; GC/MS = gas chromatography/mass
  spectrometry may also be used.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-31

-------
Method TO-4A
                                                           Pesticides/PCBs
  Magnehlic Gauge
      0-100 in.
      Exhaust Duct
       (6 in. x 10 ft)
                              Sampling Head
                              (see Figure 3)
                                                                              Voltage Variator
                                                                                Elapsed Time
                                                                                   Meter
                                                                                7-Day Timer
         Figure 1. Typical high volume air sampler for monitoring common pesticides and PCBs.
Page 4A-32
Compendium of Methods for Toxic Organic Air Pollutants
January 1999

-------
Pesticides/PCBs                                                            Method TO-4A
January 1999       Compendium of Methods for Toxic Organic Air Pollutants         Page 4A-33

-------
Method TO-4A
                                                          Pesticides/PCBs
                                     C
            Air Flow
                    Particle Fitter
                         Particle Filter
                         Support
                          Assembled
                          Sampling
                          Module
             Air Flow
             Exhaust
                                                                   Filter Retaining Ring

                                                                   Silicone Gasket
                                               102-mm
                                              Quartz-fiber
                                              Filter
                                              Filter Support Screen
                                               Filter Holder (Part 2)


                                               Silicone  Gasket


                                               Glass Cartridge

                                               Retalnlna  Screen

                                                 Sorbent


                                               Retaining  Screen



                                               Silicone  Gasket
                                                                    Cartridge
                                                                    Holder
                                                                    (Part 1)
        Figure 2. Typical absorbent cartridge assembly for sampling common pesticides and PCBs.
Page 4A-34
Compendium of Methods for Toxic Organic Air Pollutants
January 1999

-------
Pesticides/PCBs
                                                           Method TO-4A
         Exhaust Hose
                                                              4" Diameter Pullflex
                                                              Filter and Support
                                                              PUF Adsorbent
                                                              Cartridge and Support


                                                              Quick Release Connections
                                                              for Module
                                                              Quick Release Connections
                                                              for Magnahelic Gage


                                                              Flow Control Valve
                                                                 Elapsed Time Indicator
                    Figure 3. Portable high volume air sampler developed by EPA.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-35

-------
Method TO-4A
                                                             Pesticides/PCBs
       Water In
           Soxhlet
         Extraction
         Tube and
           Thimble
                                  Water Out
                                 Allihn
                                 Condenser
                                  Flosk
         (a) Soxhlet Extraction  Apparatus
               with  Allihn  Condenser
                                                     3 Ball Macro
                                                     Synder Column
                                                                           500 mL
                                                                           Evaporator
                                                                           Flosk
                                                      10 mL
                                                      Concentrator
                                                      Tube
                                (b)  Kuderna-Danish (K-D) Evaporator
                                     with Macro Synder  Column
Disposable 6 inch
Pasteur Pipette


3

^


t
Inches


s 	 /

^vvv^,
<^^


irf "''''''%


B 1 Gram Sodium Sulfate

^ 10 Gram Silica
Gel Slurry

t ; : : : : : : : O4 — mass wool Mug
                                (c) Silica Gel Clean-up Column
                     Figure 4. Apparatus used for sample clean-up and extraction.
Page 4A-36
Compendium of Methods for Toxic Organic Air Pollutants        January 1999

-------
Pesticides/PCBs
                                                        Method TO-4A
           Glass PUF Cartridge with
           Stainless Steel Screens
                   ,64mmO.D.1
                   h——	H
                                                        End Cap
               O4
               O3
               •«*
               O

               LO
           Glass
           Cartridge
                                            \   f fff. f$ •
                                                       •PUF Plug
                                                       -End Cap
                        5a. Glass PUF cartridge, plug, and end caps.
                    Accessories
             Teflon Sealing Caps   pUF Insert
               with 0-rings for
             capping PUF Sampler

                                                 Aluminum Canister for Shipping
                                                and Storage of the PUF Sampler
                             5b. PUF shipping container.
                    Figure 5.  Glass PUF cartridge (5 a) and shipping container
                         (5b) for use with high-volume sampling systems.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-37

-------
Method TO-4A                                                            Pesticides/PCBs
Page 4A-38         Compendium of Methods for Toxic Organic Air Pollutants       January 1999

-------
Pesticides/PCBs
                                                            Method TO-4A
        Mercury
       Manometer
                                                                        Barometer
                                                                              Thermometer
                                                                          Filter Adapter
                                                                          Rootsmeter
                                                                       High Volume  Motor
           Figure 6. Positive displacement rootsmeter used to calibrate orifice transfer standard.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-39

-------
Method TO-4A
                                                         Pesticides/PCBs
   <
   6
    w o
      i
    uS
      o
              00
Orifice No.
Rootsmeter No
                            05
                            ^   8'
                             i!   I
                            •53 S 5
         & 8 8 &
                                 is
                               03 >   ^
                               oi
                                                00
                                                   00
                                                   en
                                                             8
                                                            8
                                                              .a
                                                             60
                                                             a
                                                             :a
                                         a
                                         oo
                                         o
                                         o.
                                                                     O
                                                                     
                                                                     O
Calculation Equation
                                                                                  oo
                                                                                 a
                                                           II  II    H
                                                           "2 "2   "H
                                                           HV"   C?

                                                                  (S
Page 4A-40
Compendium of Methods for Toxic Organic Air Pollutants
January 1999

-------
Pesticides/PCBs
                                                           Method TO-4A
                Sampling head
                                       Calibrated
                                         orifice
                                                              Shutoff valves
                     Magnehelte gauge
                        0-100 in.
                                                                        Manometer
                                                                         0-18 in.
                                                       Pipe fitting (1/2 in.)
                                            r, , ,,,irrr.-. rfrrrrrr. ,-rrrrrrrr, .
                                                                    Elapsed time
                                                                       meter
    Exhaust duct
                 Figure 8. Field calibration configuration of the high-volume sampler
                                 for common pesticides and PCBs.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-41

-------
Method TO-4A                                                            Pesticides/PCBs
Page 4A-42         Compendium of Methods for Toxic Organic Air Pollutants       January 1999

-------
Pesticides/PCBs
                                                           Method TO-4A
                                COMPENDIUM METHOD TO-4A
               FIELD CALIBRATION DATA SHEET FOR SAMPLER CALIBRATION
Sampler ID:
Sampler Location:
High Volume Transfer Orifice Data:
  Correlation Coefficient  (CC1):
                        (CC2):
  Intercept (Bl):
          (B2):

Calibration Date:	Time:
Calibration Ambient Temperature: 	°F
Calibration Ambient Barometric Pressure:
Calibration set point (SP):  	
                              Calibration Orifice ID:
                              Job No.:

                              Slope (Ml):
                                    (M2):
                                            CALIBRATOR'S SIGNATURE
                     "Hg     mm Hg
                       o ^^—^^—      o
                                   SAMPLER CALIBRATION
Actual values from calibration
Orifice
manometer,
inches
(Yl)







Monitor
Magnehelic,
inches
(Y2)
70
60
50
40
30
20
10
Calibrated values
Orifice
manometer
(Y3)







Monitor
Magnehelic
(Y4)







Calculated value
orifice flow, scm
(XI)







                                           Definitions
Yl   = Calibration orifice reading, in. H2O
Y2   = Monitor Magnehelic reading, in. H2O
Pa   = Barometric pressure actual, mm Hg
Bl   = Manfacturer's Calibration orifice Intercept
Ml   = Manufacturer's Calibration orifice manometer
       slope
Y3   = Calculated value for orifice manometer
     = [Yl(Pa/760)(298/{Ta + 273})]'/2
                              Y4   =  Calculated value for Magnehelic
                                   =  [Y2(Pa/760)(298/{Ta + 273})]'/2
                              XI   =  Calculated value orifice flow, scm
                                       Y3-B1
                                       Ml
                              Pstd   =  Barometric pressure standard, 760 mm Hg
                              Ta   =  Temperature actual, ° C
                              Tstd   =  Temperature standard, 25° C
                    Figure 9. Orifice transfer standard field calibration data sheet.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-43

-------
Method TO-4A
                                    Pesticides/PCBs
     Y4
  "H2O adj.
             Exhaust duct
^s /
/ Motor
^*S support
1
otf—
7-
-------
Pesticides/PCBs
                                                         Method TO-4A
  Sampler I.D. No.:
  Lab PUF Sample No.:
  Sample location:
                             COMPENDIUM METHOD TO-4A
                                FIELD TEST DATA SHEET
                                GENERAL INFORMATION
                               Operator:
                               Other:
  PUF Cartridge Certification Date:
  Date/Time PUF Cartridge Installed:
  Elapsed Timer:
    Start	
    Stop 	
    Diff.	
  Sampling
                                                    Start
                            Barometric pressure ("Hg)  	
                            Ambient Temperature (°F)  	
                            Rain                    Yes
                                 Stop
                                                    No
                                Yes
                               "NO
  Ml
  M2
      Bl
      B2
Sampling time
  Start 	
  Stop 	
  Diff.	
                                                Audit flow check within ±10 of set point
                                                	Yes
                                                     No
TIME




Avg.
TEMP





BAROMETRIC
PRESSURE





MAGNEHELIC
READING





CALCULATED
FLOW RATE
(scmm)





READ BY





  Comments
                                Figure 11.  Field test data sheet.
January 1999
Compendium of Methods for Toxic Organic Air Pollutants
                                Page 4A-45

-------
Method TO-4A                                                            Pesticides/PCBs
Page 4A-46         Compendium of Methods for Toxic Organic Air Pollutants       January 1999

-------
Pesticides/PCBs
                                                              Method TO-4A
     OPERATING CONDITIONS
     Column Type:
DB-5 0.32  copillory,
0.25 um  film thickness
                                                               Dibutylchlorendate
     Column Temperature Program:   90°C(4min)/16*C per min to
                    154'C/44C  per min  to 270*C.
     Detector:        Electron Capture
     Carrier Gas:     Helium  at  1 mL/min.
     Make  Up  Gas:   5% Methane/95% Argon  at  60 mL/min.
                                                                 Methoxychlor
                             Heptachlor
                      Lindane
                                                       Endrin
                                                                p,p'DDT
                                                   Dieldrin
     TIME
         Figure 12. Chromatogram showing a mixture of single component pesticides determined
                                by GC/ECD using a capillary column.
January 1999
 Compendium of Methods for Toxic Organic Air Pollutants
Page 4A-47

-------
Method TO-4A                                                            Pesticides/PCBs
Page 4A-48         Compendium of Methods for Toxic Organic Air Pollutants       January 1999

-------
Pesticides/PCBs                                                               Method TO-4A
                                 This Page Intentionally Left Blank
January 1999        Compendium of Methods for Toxic Organic Air Pollutants          Page 4A-49

-------