EPA/500/A^03/267 Evaluation of Portable Gas Chromatographs Richard E. Berkley US Environmental Protection Agency Research Triangle Park, NC 27711 Michael Miller IIT Research Institute Chicago, IL 60616 Joseph C. Chang Battelle Columbus Laboratories Columbus, OH 43201 Karen Oliver, Christopher Fortune, and Jeffrey Adams ManTech Environmental Services Research Triangle Park, NC 27709 ABSTRACT Limits of detection, linearity of responses, and stability of response factors and retention times for five commercially-available portable gas chromatographs (PGC) were determined during laboratory evaluation. The PGCs were also operated at the French Limited Superfund site near Houston, TX during startup of bioremediation. Concentrations of volatile organic compounds (VOC) at the site were slightly above ambient background levels. Concurrent collocated grab samples were collected periodically in canisters and analyzed by Method TO-14 using a mass-selective detector. Canister data were taken to indicate correct concentrations and were used to assess the accuracy of PGC data. Durability, reliability, and complexity of operation of PGCs were also evaluated. The principal goal of the study was to determine the best way to use each instrument as a monitor for airborne VOCs. This paper has been reviewed in accordance with the U.S. Environ- mental Protection Agency's peer and administrative review policies and approved for presentation and publication. Mention of trade names or commercial products does not constitute endorsement or recommendation for use. INTRODUCTION Analysis of volatile organic compounds (VOC) in ambient air is usually done by collecting samples at field sites and returning them to a laboratory [1]. Laboratory analysis can be done with state-of- the-art procedures, but data are delayed and sample integrity could become compromised during collection or storage. A portable gas chromatograph (PGC) can produce data immediately and can deliver more information at less cost than laboratory-based methods [2]. We have evaluated five commercially-available PGCs as monitors for VOC in air. During the evaluation a field test was conducted at the French Limited Superfund Site near Crosby, TX in January 1992. ------- EXPERIMENTAL PGCs were borrowed from manufacturers. They were evaluated in a laboratory and then tested at the field site. The laboratory evaluation was repeated for some instruments because they were modified at the field site by factory personnel. Portable Gas Chromatographs All PGCs were microprocessor-controlled instruments capable of unattended autosampling. Specific features and capabilities of each PGC are summarized in Table 1. Information specific to individual instruments, which was not included in Table 1 is given below. The MSI 301 is an organic vapor monitor which operates on the principle of a gas chromatograph. It can be adjusted to detect any three compounds from a large group of potential analytes. The SRI 8610 is a field-deployable laboratory chromatograph. It can accommodate up to three detectors operating simultaneously. Other available detectors are thermal conductivity, flame ionization, nitrogen-phosphorus, thermionic ionization, electron capture, and flame photometric. Carrier gas flow is regulated by a digital flow controller, with which preselected flow rates can be set. Standards Calibration standards were prepared by dynamic dilution of commercially-prepared 10 ppm mixtures of analytes in nitrogen (Alfagaz, Scott) using humid zero air. Laboratory Evaluation Laboratory evaluation consisted of 1) learning how to operate each unit as a monitor for VOCs in air, 2) calibrating it with standard mixtures of VOCs in zero air, 3) evaluating the calibration plots, and 4) calculating detection limits from calibration data. Each unit was first set up and operated according to manufacturer's specifications. After operating procedures had been learned and proper function of the instrument had been verified, calibration was performed by analyzing standard mixtures at five levels. Reliability, durability, convenience of operation, and clarity and relevance of the operator's manual were also considered. Field Evaluation The instruments were shipped to the hazardous waste site and operated there. Data from the PGCs were compared with data obtained from concurrent collocated whole-air grab samples collected in evacuated passivated canisters. Accuracy, precision, ease of operation, and reliability were determined. Field Site The French Limited Superfund site was an abandoned flooded sand pit into which refinery waste had been dumped. Before remediation, twenty-five feet of water overlay ten feet of sludge covering an area of seven acres. The water above the sludge was clean enough to support aquatic life, but volatile solvents were leaching from the sludge into the surface aquifer. Bioremediation at the site was performed using selectively-nurtured indigenous microorganisms. Pond fluid was pumped out of the pond, injected with oxygen gas and nutrients, then pumped back into the pond. Dredges loosened sludge from the bottom and high-speed stirrers mixed it with the water. The ------- bacteria fed on the sludge, converting it into biomass which eventually decayed into carbon dioxide and water. Agitation of the pond resulted in a slight increase in atmospheric concentrations of organic vapors due to release of solvents dissolved in the sludge. Concentrations were lower than anticipated from a pilot study in which air had been used to provide oxygen. Diesel exhaust from dredges and agitators contributed significantly to airborne VOC levels. The evaluation was conducted at a power-control shed which was located 20 feet south and 15 feet above the edge of the pond. Initial attempts to conduct the study on an outdoor deck were defeated by inclement weather, and all units were moved inside the shed where temperature was maintained at 20°C. All units were connected to 110 volt AC power. Each unit used its own sample pump to import outside air through 1/B inch OD stainless steel tubing. Calibrations were performed periodically. Grab sampling was done by opening the valve of an evacuated passivated canister while holding it within three feet of the ends of the intake tubes as the PGCs sampled simultaneously. Grab samples were returned to the EPA laboratory in Research Triangle Park where they were analyzed by GC/MSD according to Method TO-14. Canister data were assumed to accurately reflect concentrations of compounds in the air sampled by the PGCs. RESULTS AND DISCUSSION The ability of an instrument to reliably analyze VOCs in ambient air depends upon stability of retention times, adequate chromatographic resolution, and the sensitivity, selectivity, and stability of the detector. Stable performance is a significant problem for any sensitive instrument which travels, and it was one of two principal subjects of investigation in this study. The other subject of investigation was accuracy, which was determined by comparing PGC data with data from simultaneous canister grab samples. Retention Time Stability Retention time stability is crucial to accurate identification of compounds. It was assessed in terms of the standard deviations of retention times. Retention times from each calibrated compound were averaged, and the standard deviation was calculated. Average retention times are shown as horizontal bars in Figure 1, and standard deviations are indicated at the end of each bar. They were usually small but discernible. During the Scentograph field calibration, the carrier flow rate was changed, producing two distinct groups of retention times and a large standard deviation. This was probably the result of human error and not due to a fault in the instrument. Standard deviations in all other cases were small, but they needed to be even smaller. The importance of retention time stability to compound identification cannot be over-emphasized. It is vital to be able to set carrier flow quickly and accurately. Column temperature stability is even more important, because a shift in temperature changes relative retention times and invalidates the entire calibration table. Ideally, the standard deviations in Fig 1 should have been almost invisible at the scale of the drawing. Detector Stability Stable detector response is crucial to accurate quantitation. It was evaluated by observing the scatter in the calibration plots (Figures 2-4) and by noting how closely the correlation coefficients of the plots approached unity (Table 2). Significant scatter was ------- taken to indicate an unstable response or possibly random sample loss. The calibration plots showed some scatter, but all of the instruments were able to report approximately correct data within the range calibrated. In several cases precision was mediocre. Sensitivity. Sensitivity is indicated by the detection limit (Table 3), which was taken to be the concentration corresponding to a response equal to four times the baseline noise adjacent to the peak. Baseline noise was difficult to evaluate in some cases. Neither the Scentograph nor the Photovac sent a real baseline signal to external devices, and the baseline of the MSI 301 was difficult to locate. Estimates of baseline noise were made by visual inspection in those cases. The lower the correlation coefficient, the less credible are calculated detection limits. Detection limits were usually higher for the field calibrations, reflecting the stress of field operation. Resolution and Selectivity Resolution of PGCs is generally lower than for laboratory instruments because 1) shorter columns are used to get quicker results, 2) autosampling schemes tend to introduce dead volume into the system and thus broaden injection bandwidths, 3) field operation tends to degrade performance because instruments are vulnerable to vibration, rough handling, dust, and fluctuating ambient temperatures. Even the best resolution achievable cannot separate all of the compounds present at background levels in ambient air. Typical chromatograms obtained in laboratory analyses display as many as 200 peaks with up to four compounds coeluting in each peak. Detector selectivity must be considered together with resolution when assessing the ability of an instrument to distinguish target compounds from other compounds in authentic field samples. To some extent, selectivity substitutes for resolution in a PGC. The resolution of toluene from benzene was calculated for each PGC in the study to be twice the difference in retention times divided by the sum of the peak widths at base height. These figures are listed in Table 4 with a summary of the compounds to which each detector responds. Field Performance Ambient levels of analytes found during the field study were scarcely above calculated detection limits, but PGC results generally resembled data from the corresponding grab sample. Some discrepancies were probably caused by poor mixing of air close downwind of the source. Agreement between methods was not exact, but it was close enough to show that all of the PGCs provided reasonable estimates of the concentrations of compounds for which they were calibrated. In Table 5 these data are summarized by the average of the averages for the individual compounds and the average of the standard deviations for the individual compounds. The ranges shown in Table 5 are the difference between the best and worst agreement between paired canister and PGC results among all of the compounds. These data are a further condensation of data which were tabulated previously [3]. A small average difference with a smaller standard deviation and a narrow range would indicate close agreement between the two methods. A large average difference with a small standard deviation would indicate good precision but substantial disagreement due to systematic error. A small average difference with a standard deviation of about the same magnitude as the average difference would indicate data of reasonable accuracy but mediocre precision, which would be expected ------- when analyzing concentrations near detection limits. That is generally what is seen in Table 5. A somewhat larger average difference and a standard deviation larger than the average difference with a broad range indicates some outliers among the data. A very large average difference with a standard deviation as large as the range would indicate little or no agreement between methods, and that is not seen in Table 5. CONCLUSIONS All of the PGCs performed substantially as expected. It would be futile to try to rank them on the basis of performance, since they are not uniformly applicable to the same compounds. All instruments evaluated in this study performed substantially as claimed. Each was able to detect compounds at the levels encountered, generally with reasonable accuracy. A variety of instruments and detectors may be required to match capability with need. Choice of an instrument for a particular application should be based upon consideration of its particular features and capabilities. All of the instruments in this study were seriously handicapped by exposure to cold wind, and they all required shelter in order to function properly. REFERENCES 1. Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air. Environmental Protection Agency, Atmospheric Research and Exposure Assessment Laboratory, Research Triangle Park, NC 27711. EPA-600/4-84-017. June 1988. 2. R. E. Berkley, J. L. Varns, and J. Pleil. Comparison of Portable Gas Chromatographs and Passivated Canisters for Field Sampling Airborne Toxic Organic Vapors in the United States and the USSR. Environ. Sci. Technol., 25(8): 1439 (1991). 3. R. E. Berkley, M. Miller, J. C. Chang, K. Oliver, C. Fortune, Evaluation of Commercially-Available Portable Gas Chromatographs, in Proc. 1992 EPA/AWMA International Symposium on Measurement of Toxic and Related Air Pollutants, Air and Waste Management Association, Pittsburgh, PA, 1992, pp 413-418. 4. R. E. Berkley, K. Kroraniller, and K. Oliver, Performance Optimization of Photovac 10S70 Portable Gas Chromatograph. Proceedings of the 1990 EPA/AWMA International Symposium: Measurement of Toxic and Related Air Pollutants, 1990 pp 849-854. ------- TABLE 1. SUMMARY OF INSTRUMENT FEATURES AND CAPABILITIES Power Column °C HNU 311 110 V AC isothermal to 200 Detector PID Injector injection loop or syringe port Back- flush yes MSI 301 110 V.AC 12 V DC isothermal 65 SAW precon- centrate & desorb no Photovac 10S+ 110 V AC 12 V DC isothermal 40, 50 PID injection loop or syringe port yes Scento- graph 110 V AC 12 V DC isothermal to 180 AID precon- centrate & desorb or syringe port no SRI 8610 110 V AC 12 V DC program to 250 PID + ELCD precon- centrate & desorb or syringe port no Carrier Data report helium, & others printer or computer scrubbed ambient air internal processor, external computer, or printer ultra- zero air magnetic card, external printer, or computer argon helium computer disk helium, nitrogen, scrubbed ambient air external printer or computer PID photoionization detector SAW surface acoustic wave detector TCD thermal conductivity detector ECD electron capture detector AID argon ionization detector ELCD electrolytic conductivity detector ------- TABLE 2. CORRELATION COEFFICIENTS OF CALIBRATION PLOTS Tetra- chloro- Chloro- Benzene Toluene ethylene benzene o-Xylene HNU 311 Lab 0.958 0.958 0.976 0.977 0.996 Field 0.995 0.993 0.996 0.976 0.966 MSI 301 Lab 0.985 0.964 NA NA 0.922 Field 0.971 0.971 NA NA 0.979 Photovac 10S+ Lab 0.997 0.998 0.999 0.997 0.996 Field 0.994 0.996 0.999 0.990 0.988 Scentograph Lab 0.939 0.836 0.955 0.817 ND Field 0.964 0.954 0.992 0.870 0.919 SRI 8610 PID Lab 0.990 0.991 0.993 0.981 0.977 Field 0.963 0.955 0.977 0.959 0.972 SRI 8610 ELCD Lab NA NA 0.965 0.903 NA Field NA NA 0.845 0.850 NA TABLE 3. DETECTION LIMITS CALCULATED FROM CALIBRATION PLOTS (parts per billion by volume) Tetra- chloro- Chloro- Benzene Toluene ethylene benzene o-Xylene HNU 311 Lab 1.9 2.4 3.1 2.0 5.3 Field 2.5 3.2 4.6 4.1 9.3 MSI 301 Lab 2.4 9.1 NA NA 4.5 Field 9.0 10.3 NA NA 10.1 Photovac 10S+ Lab 1.1 2.6 2.1 3.6 14.4 Field 0.9 2.0 1.3 3.3 20.6 Scentograph Lab 5.1 4.4 5.9 5.5 27.9 Field 6.7 7.5 3.4 6.6 46.6 SRI 8610 PID Lab 3.4 2.7 2.9 2.0 1.1 Field 4.9 3.4 3.4 2.2 3.0 SRI 8610 ELCD Lab NA NA 1.9 3.5 NA Field NA NA 2.2 8.8 NA ND Not detected. NA Detector does not respond. ------- TABLE 4. CHROMATOGRAPHIC RESOLUTION AND DETECTOR SELECTIVITY Resolution of benzene from toluene was taken to be twice the difference in retention times divided by the sum of the peak widths at base height. HNU 311 MSI 301 Photovac 10+ Scentograph SRI 8610 SRI 8610 Resolution 2.58 1.09 7.77 7.23 5.32 5.32 Detector PID SAW PID AID PID ELCD Compounds Detected Aromatics, halocarbons Benzene, toluene, xylene Aromatics, halocarbons Universal Aromatics, halocarbons Halocarbons TABLE 5. AVERAGE ABSOLUTE DIFFERENCES BETWEEN TO-14 REFERENCE METHOD AND PORTABLE GAS CHROMATOGRAPH DATA Average differences between paired canister and PGC results for the five compounds were averaged to summarize performance of the instrument. Standard deviations for individual compounds were also averaged. Ranges are the differences between the largest and smallest discrepancies among all five compounds. HNU 311 15 Runs, 5 Compounds MSI 301 13 Runs, 2 Compounds Photovac 10+ 14 Runs, 5 Compounds Scentograph 13 Runs, 5 Compounds SRI 8610 PID 13 Runs, 5 Compounds Average of per Compound Average Differences 4.80 1.75 0.63 1.08 1.18 Average of per Compound Standard Deviations 15.13 0.60 0.73 1.00 Maximum Range of All Differences 61.15 7.20 2.20 2.58 3.98 SRI 8610 ELCD 13 Runs, 2 Compounds 1.30 5.30 ------- HNU 311 PID Lab HNU 31 I PID Field PV 105* PID Lab Benzene| Toluene TetroehIoroethy11 Ch1orobeazene o"Xyier*e PY 10S+ PID Field ScenLoar-aph Lab AID Scentaaraph Field AID SRI 8610 PID Lob SRI 8610 PID Field SRI 8610 ELCD Lab SRI 8610 ELCD Field MSI 301 SAW Lob MSI 301 SAW Field ea s CD S3 a Retent 1 on TI m © Figure 1. Retention time stabilities. Standard deviation of retention time for each compound is represented by a shaded bar centered on the end of an unshaded bar which represents average retention time. ------- MSI 301 Calibrations 5000 5000 Q) 0) 500 , 2500 ~ ~_ 0) (D Lab •fc. es EJ S Scentograph Ca I 1 brat i oris 2000000 1000000 C ~ CL tn Q) cr 4- O S3 U1 ca s GJ 10000000 5000000. Concentration (ppbv) Benzene Toluene x Tetrochloroethy1 en I i i D LhIorobenzene Concentret1 on (ppbv) ° o-Xylene Figure 2. Calibration plots from laboratory and field studies for MSI 301 and Sentex Scentograph. ------- HNU 31 1 Col i br at. 1 ans M 200000 D-1000B0 Field Lab in c Q CL in Q) cr + o Ph olovac 10S+ c o 1 i brat i oris 100. . 100 Concent-notion ( ppbv ) Benzene * To 1uene D 50. Field Tetrechlonoethylene Chi orobenzene Concentroiion (ppbv) o - Xy 1 ene Figure 3. Calibration plots from laboratory and field studies for HNU 311 and Photovac 10S+. ------- 1000 SRI 8610 RID Colibrotlon: 1000. 0) CL 0) 0) CZ 500. 500 s Q Tsj O S O) S es A s> o Field SRI 8610 ELCD Calibrations 4000 3000 Concentration (ppbv) Concentration (ppbv) + Benzene x TeLrach1oroethy1ene o o ~Xy1ene ° Toluene 0 Chlorobenzene Figure 4. Calibration plots from laboratory and field studies for SRI 8610 photoionization and electrolytic conductivity detectors. ------- technical report data (Pleeit read Imtruttions on the rtvtnt be fort eompktin, 1. REPORT NO. 3. EPA/600/A-9S/2G7 4. TITLE ANO SUBTITLE Evaluation of Portable Gas Chromatographs S. REPORT DATE B. PERFORMING ORGANIZATION CODE 7. AUTHOR(S) R. E. Berkley, M. Miller, J. Chang, K. Oliver, C. Fortune, J. Adams B. PERFORMING ORGANIZATION REPORT NO. 9. PERFORMING ORGANIZATION NAME ANO ADDRESS IIT Research Institute ManTech Environmental Serv. 10 West 35 Street 2 Triangle Drive Chicago, IL 60616 RTF, NC 27709 10. PROGRAM ELEMENT NO. Y105C/A04/F01 11. CONTRACT/GRANT NO. 68-D8-0002 68-D0-0106 12, SPONSORING AGENCY NAME AND ADDRESS AREAL/MRDD/AMRB MD-44 RTP, NC 27711 13. TYPE OF REPORT AND PERIOD COVERED Conference Proceeding 14. SPONSORING AGENCY CODE 15. SUPPLEMENTARY NOTES 16. ABSTRACT Limits of detection, linearity of responses, and stability of response factors and retention times for five commercially-available portable gas chromatographs (PGC) were determined during laboratory evaluation. The PGCs were also operated at the French Limited Superfund site near Houston, TX during startup of bioremediation. Concentrations of volatile organic compounds (VOC) at the site were slightly above ambient background levels. Concurrent collocated grab samples were collected periodically in canisters and analyzed by Method TO-14 using a mass-selective detector. Canister data were taken to indicate correct concentrations and were used to assess the accuracy of PGC data. Durability, reliability, and complexity of operation of PGCs were also evaluated. The principal goal of the study was to determine the best way to use each instrument as a monitor for airborne VOCs. 1?. KEY WORDS AND DOCUMENT ANALYSIS 1. DESCRIPTORS b.lDENTIFIERS/OPIN ENDED TERMS c. COS ATI Field/Group U. DISTRIBUTION STATEMENT 1B. SECURITY CLASS (ThisReport/ ai.NO. OF PAGES 12 30. SECURITY CLASS (This page/ 22. PRICE CPA Farm 2220-1 (R»». 4-77) RREVIOU* edition li OBtOLtTC ------- |