Rapid Radiochemical Method
Strontium-90 (90Sr) in Water Samples
EPA's rapid radiochemical methods expedite analytical
turnaround time for selected radionuclides while providing
quantitative results that meet measurement quality objectives.
Methods are applicable to samples where contamination is from
either known or unknown origins, This fact sheet is intended for
radioanalytical laboratory personnel, decision makers within the
incident command structure, additional reoccupancy decision
makers (e.g., state and local public health), and other field
environmental response personnel.
Method Summary: Strontium is isolated from the matrix and purified from potentially interfering radionuclides and matrix constituents using a
strontium-specific, rapid chemical separation method. The sample is equilibrated with strontium carrier, and concentrated by strontium/barium
carbonate coprecipitation. If insoluble residues are noted during acid dissolution steps, the residue and precipitate mixture is digested in nitric
acid to solubilize strontium. The solution is passed through a strontium specific resin extraction chromatography column. The sample test
source is promptly counted on a gas flow proportional counter to determine the beta emission rate, which is used to calculate the total
radiostrontium activity.
Time to Process: 8.7 hr
Includes radiochemical processing and counting
Compare to traditional method (EPA 905.0): 25-28 hr
Measurement Quality Objectives
Required method uncertainty: 1.0 pCi/L
Analytical action level (AAL): 8 pCi/L
Required relative uncertainty: 13% above ML
Minimum detectable concentration: 1.5 pCi/L
Sample quantity: ~ 500 ml_
Count time: ~ 1.25 hr
Sample Preservation
Samples should be collected in 1 L plastic containers
Analysis within 3 days of sampling: No preservation required
Holding time >3 days: Adjust pH to <2 with concentrated nitric
acid
Waste Generated per Sample
Nitric acid
Hydrochloric acid
mg/mL levels of barium in initial column effluents
Radiostrontium contained in final precipitated materials
Used resins and columns should be considered radioactive waste
Method Access: https://www.epa.gov/sites/production/files/2015-
06/documents/sr-90 in water rev 0 1 epa 402-r-10~001d.pdf
Method Application
The method provides a very rapid non-radioisotope-
specific screen for total radiostrontium in drinking water
and other aqueous samples. Application of this method
should be validated by the laboratory using the
protocols provided in Method Validation Guide for
Qualifying Methods Used by Radiological Laboratories
Participating in Incident Response Activities, or the
protocols published by a recognized standards
organization for method validation.
Equipment and Supplies
Analytical balance: 10~4 g readability or better |
Centrifuge able to accommodate 250 mL flasks and
50 mL centrifuge tubes | Centrifuge flasks: 250 mL,
disposable | Centrifuge tubes: 50 mL, disposable |
Low-background gas flow proportional counter |
Laboratory supplies: pH paper; stainless steel
planchets or other sample mounts, ~ 2 inch diameter
| Vacuum system: box; pump or laboratory system;
white inner tips; yellow outer tips