-------
time since the rates can vary depending on how the toxicant affects the bacteria. Since these
samples were all treated as unknowns, both data collection times were used.
For each contaminant, Microtox® analyzed the lethal dose concentration and three additional
concentration levels four times. Only one concentration of potential interference was analyzed.
The absolute light units were recorded and the percent inhibition was calculated for each
sample. When Microtox® produced percent inhibitions greater than 50% for a contaminant, EC50
(effective concentration causing 50% inhibition) values were also calculated and reported. Two
operators performed all the analyses using Microtox®. Both held bachelor's degrees in the
sciences and spent approximately four hours with the vendor to become familiar with using
Microtox®.
3.4.3 Stock Solution Confirmation Analysis
The concentrations of the contaminant and interfering compound stock solutions were verified
with standard analytical methods, with the exception of colchicine, ricin, and botulinum
toxin—contaminants without standard analytical methods. Aliquots to be analyzed by standard
methods were preserved as prescribed by the method. In addition, the same standard methods
were used to measure the concentrations of each contaminant/potential interference in the
unspiked DDW so that background concentrations of contaminants or potential interferences
were accounted for within the displayed concentration of each contaminant/potential
interference sample. Table 3-3 lists the standard methods used to measure each analyte; the
results from the stock solution confirmation analyses (obtained by reporting the lethal dose
concentration for the contaminants and the single concentration that was analyzed for the
potential interferences); and the background levels of the contaminants and potential inter-
ferences measured in the DDW sample, which were all non-detect or negligible.
Standard methods were also used to characterize several water quality parameters such as the
concentration of trihalomethanes, haloacetic acids, and total organic halides; turbidity; dissolved
organic carbon content; pH; alkalinity; specific conductivity; and hardness. Table 3-4 lists these
measured water quality parameters for both the water sample collected in Columbus, Ohio,
representing a water system using chlorination as the disinfecting process, and the water sample
collected in St. Petersburg, Florida, representing a water system using chloramination as the
disinfecting process.
9
-------
Table 3-3. Dose Confirmation Results
Method
Average Concentration ±
Standard Deviation N = 4
(mg/L)
Background
in DDW
(mg/L)
Contaminant
Aldicarb
EPA 531.1(3)
280 ± 28
<0.0007
Colchicine
(a)
NA®
NA
Cyanide
EPA 335.1(4)
250 ± 15
0.008
Dicrotophos
EPA SW846 (8141A)®
1,400 ± 140
<0.002
Thallium sulfate
EPA 200.8(6)
2,400 ± 24
<0.001
Botulinum toxin
(a)
NA
NA
Ricin
(a)
NA
NA
Soman
(<0
0.068(d) ± 0.001
<0.05
VX
(<0
0.20 ± 0.02
<0.05
Potential
Interference
Aluminum
EPA 200.8
0.36 ±0.01
<0.10
Copper
EPA 200.8
0.65 ±0.01
0.011
Iron
EPA 200.8
0.069 ± 0.008
<0.04
Manganese
EPA 200.8
0.26 ±0.01
<0.01
Zinc
EPA 200.8
3.5 ±0.35
0.3
(a) No standard method available. QA audits and balance calibration assured accurately prepared solutions.
(b) NA = Not applicable.
(c) Purity analyses performed on chemical and biological agent materials using Battelle standard operating procedures.
(d) The result of the dose confirmation analysis for soman was 23% of the expected concentration of 0.30 mg/L.
10
-------
Table 3-4. Water Quality Parameters
Parameter
Method
Dechlorinated Columbus,
Ohio, Tap Water (disinfected
by chlorination)
Dechlorinated
St. Petersburg, Florida,
Tap Water (disinfected by
chloramination)
Turbidity
EPA 180.1(7)
0.1 NTU(a)
0.3 NTU
Organic carbon
SM 5310(8)
2.5 mg/L
2.9 mg/L
Specific conductivity
SM 2510(S)
364 |imho
460 |imho
Alkalinity
SM 2320(S)
42 mg/L
97 mg/L
pH
EPA 150.1(9)
7.65
7.95
Hardness
EPA 130.2(9)
112 mg/L
160 mg/L
Total organic halides
SM 5320B(8)
190 [ig/L
83 jig/L
Total trihalomethanes
EPA 524.2(10)
52.8 jig/L
2.4 |ig/L
Total haloacetic acids
EPA 552.2(11)
75.7 fig/L
13.5 fig/L
(a)NTU = nephelometric turbidity unit
11
-------
Chapter 4
Quality Assurance/Quality Control
QA)/QC procedures were performed in accordance with the quality management plan (QMP) for
the AMS Center(12) and the test/QA plan for this verification test.(1)
4.1 Quality Control of Stock Solution Confirmation Methods
The stock solutions for aldicarb, cyanide, dicrotophos, and thallium sulfate were analyzed using
a standard reference method at ATEL. As part of ATEL's standard operating procedures (SOPs)
various QC samples were analyzed with each sample set. These included matrix spike,
laboratory control spike, and method blank samples. According to the standard methods used for
the analyses, recoveries of the QC spike samples analyzed with samples from this verification
test were within acceptable limits of 75% to 125%, and the method blank samples were below
the detectable levels for each analyte. For VX and soman, the confirmation analyses were
performed at Battelle using a Battelle SOP. Calibration standard recoveries of VX and soman
were always between 69% and 130%, and most of the time were between 90% and 100%.
Standard analytical methods for colchicine, ricin, and botulinum toxin were not available and,
therefore, not performed. QA audits and balance calibrations assured that solutions for these
compounds were accurately prepared.
4.2 Quality Control of Drinking Water Samples
A method blank sample consisting of ASTM Type IIDI water was analyzed once by Microtox®
for approximately every 20 drinking water samples that were analyzed. No detectable inhibition
was observed in any of these samples. A positive control sample also was analyzed once for
approximately every 20 drinking water samples. While performance limits were not placed on
the results of the positive control sample, the vendor informed Battelle that, if the positive
control samples did not cause greater than approximately 50% inhibition, it would indicate to
the operator that Microtox® was operating incorrectly. For nine positive control samples of
phenol, inhibitions of 86% ±1% and 87% ± 2% were measured at 5 and 15 minutes,
respectively. For 17 positive control samples of zinc sulfate, inhibitions of 77% ± 1% and 96%
± 2% were measured at 5 and 15 minutes. These inhibition values indicated the proper
functioning of Microtox®. A negative control sample (unspiked DDW) was analyzed with
approximately every four samples. The percent inhibition calculation incorporated the inhibition
of the negative control; therefore, by definition, the negative control samples had 0% inhibition.
12
-------
4.3 Audits
4.3.1 Performance Evaluation Audit
The concentration of the standards used to prepare stock solutions of the contaminants and
potential interferences was confirmed by analyzing solutions of each analyte prepared in ASTM
Type IIDI water from two separate commercial vendors using the confirmation methods. The
standards from one source were used to prepare the stock solutions during the verification test,
while the standards from a second source were used exclusively to confirm the accuracy of the
measured concentration of the first source. The percent difference (%D) between the measured
concentration of the performance evaluation (PE) sample, and the prepared concentration of that
sample was calculated using the following equation:
%D=—Xl00%
A (1)
where M is the absolute value of the difference between the measured and the prepared concen-
tration, and A is the prepared concentration. The %D between the measured concentration of the
PE standard and the prepared concentration had to be less than 25% for the measurements to be
considered acceptable. Table 4-1 shows the results of the PE audit for each compound. All %D
values were less than 25.
Table 4-1. Summary of Performance Evaluation Audit
Average Measured
Concentration ±
Standard Deviation
(mg/L)
Actual Concentration
(mg/L)
Percent
Difference
Aldicarb
0.00448 ± 0.000320
0.00500
11
Contaminant
Cyanide
Dicrotophos
0.207 ± 0.026
0.00728 ± 0.000699
0.200
0.00748
4
3
Thallium sulfate
0.090 ± 0.004
0.100
10
Potential
interference
Aluminum
Copper
0.512 ±0.013
0.106 ±0.002
0.500
0.100
2
6
Iron
0.399 ± 0.004
0.400
0.30
Manganese
0.079 ± 0.003
0.100
21
Zinc
0.106 ±0.016
0.100
6
Given the lack of confirmation methodology for some of the contaminants in this verification
test, PE audits were not performed for all of the contaminants. PE audits were performed when
13
-------
more than one source of the contaminant or potential interference was commercially available
and when methods were available to perform the confirmation. To assure the purity of the other
standards, documentation, such as certificates of analysis, was obtained for colchicine,
botulinum toxin, and ricin. In the cases of VX and soman, which were obtained from the U.S.
Army, the reputation of the source, combined with the confirmation analysis data, provided
assurance of the concentration analyzed.
4.3.2 Technical Systems Audit
The Battelle Quality Manager conducted a technical systems audit (TSA) to ensure that the
verification test was performed in accordance with the test/QA plan(1) and the AMS Center
QMP.(12) As part of the audit, the Battelle Quality Manager reviewed the contaminant standard
and stock solution confirmation methods, compared actual test procedures with those specified
in the test/QA plan, and reviewed data acquisition and handling procedures. Observations and
findings from this audit were documented and submitted to the Battelle verification test
coordinator for response. No findings were documented that required any significant action. The
records concerning the TSA are permanently stored with the Battelle Quality Manager.
The EPA Quality Manager also conducted a TSA to ensure that the verification test was
performed in accordance with the test/QA plan(1) and the AMS Center QMP.(12) As part of the
audit, the EPA Quality Manager compared actual test procedures with those specified in the
test/QA plan and reviewed data acquisition and sample preparation records and procedures. No
significant findings were observed during the EPA TSA. The records concerning the TSA are
permanently stored with the EPA Quality Manager.
4.3.3 Audit of Data Quality
At least 10% of the data acquired during the verification test were audited. Battelle's Quality
Manager traced the data from the initial acquisition, through reduction and statistical analysis,
to final reporting, to ensure the integrity of the reported results. All calculations performed on
the data undergoing the audit were checked.
4.4 QA/QC Reporting
Each internal assessment and audit was documented in accordance with Sections 3.3.4 and 3.3.5
of the QMP for the ETV AMS Center.(12) Once the assessment report was prepared, the Battelle
verification test coordinator ensured that a response was provided for each adverse finding or
potential problem and implemented any necessary follow-up corrective action. The Battelle
Quality Manager ensured that follow-up corrective action was taken. The results of the TSA
were sent to the EPA.
14
-------
4.5 Data Review
Records generated in the verification test were reviewed before these records were used to
calculate, evaluate, or report verification results. Table 4-2 summarizes the types of data
recorded. The review was performed by a technical staff member involved in the verification
test, but not the staff member who originally generated the record. The person performing the
review added his/her initials and the date to a hard copy of the record being reviewed.
Table 4-2. Summary of Data Recording Process
Data to be
Recorded
Responsible
Party
Where
Recorded
How Often
Recorded
Disposition of Data(a)
Dates, times of test
events
Battelle
Laboratory
record books
Start/end of test, and
at each change of a
test parameter
Used to organize/check
test results; manually
incorporated in data
spreadsheets as
necessary
Sample preparation
(dates, procedures,
concentrations)
Battelle
Laboratory
record books
When each sample
was prepared
Used to confirm the
concentration and
integrity of the samples
analyzed
Procedures entered into
laboratory record books
Test parameters
(contaminant
concentrations,
location, etc.)
Battelle
Laboratory
record books
When set or
changed
Used to organize/check
test results, manually
incorporated in data
spreadsheets as
necessary
Stock solution
confirmation
analysis, sample
analysis, chain of
custody, and
results
Battelle or
contracted
laboratory
Laboratory
record books,
data sheets, or
data acquisition
system, as
appropriate
Throughout sample
handling and
analysis process
Transferred to
spreadsheets/agreed
upon report
(a) All activities subsequent to data recording were carried out by Battelle.
15
-------
Chapter 5
Statistical Methods and Reported Parameters
The statistical methods presented in this chapter were used to verify the performance parameters
listed in Section 3.1.
5.1 Endpoints and Precision
Microtox® reports absolute light units as a measure of light intensity (/). Each sample was com-
pared with a reference sample that, for this verification test, was unspiked DDW. This
comparison was made by calculating gamma (Gr), the ratio of the light lost at time t to the light
remaining at time t, using the two following equations:
R, -
hi (2)
r Rt 'hi i
Gt — 1
1 St
(3)
where Rt corrects for any inhibition induced by the negative control sample and Ict and Ici are the
absolute light intensities produced by the negative control at time t and at the initial time, /,
respectively. In addition, Isi and Ist are the light intensities produced by the water sample at the
initial time, i, and time t, respectively. Gt is converted to percent inhibition using the following
equation:
% inhibition = , xlOO ^
1+Gt
Percent inhibition data were calculated and are presented with respect to each test sample
analyzed as a part of this verification test. For contaminants that induced inhibition of greater
than 50%, the concentration of contaminant that affects 50% of the bacteria in the Microtox®
reagent (EC50) was estimated from the linear regression of the log of each concentration level of
the contaminant versus the percent inhibition. For contaminants that did not induce inhibition of
greater than 50%, this calculation was not appropriate.
16
-------
The standard deviation (S) of the results for the replicate samples was calculated, as follows, and
used as a measure of technology precision at each concentration.
n / -s2
X('* -')
1/2
(5)
where n is the number of replicate samples, Ik is the percent inhibition measured for the klh
sample, and I is the average percent inhibition of the replicate samples. Because the average
inhibitions were frequently near zero for this data set, relative standard deviations would often
have greatly exceeded 100%, making the results difficult to interpret. Therefore, the precision
results were left in the form of standard deviations so the reader could easily view the
uncertainty around the average for results that were both near zero and significantly larger than
zero.
5.2 Toxicity Threshold
The toxicity threshold was defined as the lowest concentration of contaminant to exhibit a
percent inhibition significantly greater than the negative control. Also, the inhibition of the
toxicity threshold had to be significantly different than the inhibition of the other concentrations
analyzed. Since the inhibition of the test samples was calculated with respect to the inhibition of
each negative control sample, the percent inhibition of the negative control was always zero. An
inhibition was significantly greater than the negative control if the average inhibition plus or
minus the standard deviation did not include zero.
5.3 False Positive/Negative Responses
A response would be considered false positive if an unspiked drinking water sample produced
an inhibition significantly greater than zero when determined with respect to ASTM Type IIDI
water. Depending on the degree of inhibition in the sample, toxicity due to subsequent contam-
ination of that sample may not be detectable or could be exaggerated as a result of the baseline
inhibition. To test for this possibility, the percent inhibition of the unspiked drinking water was
determined with respect to ASTM Type II DI water. Drinking water samples collected from
water systems using chlorination and chloramination as the disinfecting process were analyzed
in this manner. An inhibition was considered significantly different from zero if the average
inhibition, plus or minus the standard deviation, did not include zero.
A response was considered false negative when Microtox® was subjected to a lethal concentra-
tion of some contaminant in the DDW and did not indicate inhibition significantly greater than
the negative control (zero inhibition) and the other concentration levels analyzed. Requiring the
inhibition of the lethal dose sample to be significantly greater than zero and the other concentra-
tion levels more thoroughly incorporated the uncertainty of all the measurements made by
17
-------
Microtox® in determining a false negative result. A difference was considered significant if the
average inhibition plus or minus the standard deviation did not encompass the value or range of
values that were being compared.
5.4 Other Performance Factors
Ease of use (including clarity of the instruction manual, user-friendliness of software, and
overall convenience) was qualitatively assessed throughout the verification test through
observations of the operators and verification test coordinator. Sample throughput was evaluated
quantitatively based on the number of samples that could be analyzed per hour.
18
-------
Chapter 6
Test Results
6.1 Endpoints and Precision
Tables 6-la-i present the percent inhibition data for nine contaminants, and Table 6-2 presents
data for five potential interferences and the drinking water samples disinfected by both chlorina-
tion and chloramination. Given in each table are the concentrations analyzed, the percent
inhibition results for each replicate at each concentration, and the average and standard
deviation of the inhibition of the four replicates at each concentration. Results are given for the
samples analyzed at five minutes and then again at 15 minutes. For the most part, the results at
both time intervals were consistent, but according to the vendor protocol for unknown samples,
both data sets were collected and reported. EC50 values also are given when applicable. Samples
that produced negative percent inhibition values indicated an increase in light production by the
bacteria and were considered non-toxic.
6.1.1 Contaminants
The contaminants that were analyzed by Microtox® during this verification test produced one of
two trends apparent from Tables 6-la-i. Contaminants caused percent inhibitions that, starting
from the lowest concentration that produced inhibitions near zero, either increased in proportion
to the concentration in the sample so the two highest concentrations had significantly higher
inhibition, or did not change considerably, regardless of what concentration was analyzed.
Aldicarb, dicrotophos, and thallium sulfate (15 minutes only) fall into the former category, while
botulinum toxin, ricin, VX, and soman fall into the latter category. The only exceptions to these
trends were colchicine, for which the lethal dose (highest concentration) exhibited a higher
percent inhibition than for the rest of the concentration levels, and cyanide, for which the two
highest concentrations exhibited a higher percent inhibition than for the other two concentration
levels at 5 minutes. At 15 minutes, all four concentration levels of cyanide had distinct
inhibitions. For aldicarb, cyanide, and dicrotophos, whose inhibitions increased with
concentration and spanned the range from approximately no inhibition to greater than 50%
inhibition, EC50 values were calculated and reported in Tables 6-la, 6-lc, and 6-Id. Because
inhibitions did not reach 50% for the other contaminants, EC50 values could not be calculated.
19
-------
Table 6-la. Aldicarb Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
tration Inhibition Average Deviation EC50
(mg/L) (%) (%) (%) (mg/L)
Standard
Inhibition Average Deviation EC50
(%) (%) (%) (mg/L)
-3
0.28 2 3 4
6
45
-2
2 3 5
10
5
2.8 ^ 5 2
2
6
2 4 3
8
3
29
30
28 30 1
30
30
29
31 2
33 31
32
41
280 11
(Lethal ' 79 2
Dose) ^
79
83 81 2
80
Table 6-lb. Colchicine Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
tration Inhibition Average Deviation EC50
(mg/L) (%) (%) (%) (mg/L)
Standard
Inhibition Average Deviation EC50
(%) (%) (%) (mg/L)
8
0.24 ^ 5 3
6
5
2 3 2
1 3
3
NA(a)
7
2.4 ^ 6 1
6
6
KT A (a)
3
6 5 1
6
7
24 5 5 2
2
5
3
2 3
1 3
-1
240 !o
(Lethal " 11 1
Dose) j j
11
11 12 1
13
13
(a) NA = Not applicable.
20
-------
Table 6-lc. Cyanide Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen-
Standard
Standard
tration
Inhibition Average
Deviation
EC50
Inhibition
Average
Deviation
EC-50
(mg/L)
(%)
(%)
(%)
(mg/L)
(%)
(%)
(%)
(mg/L)
7
10
0.25
5
5
11
7
3
5
7
11
8
3
12
50
2.5
5
7
9
4
41
44
46
4
12
8
48
4
82
97
25
83
83
85
83
1
96
96
95
96
1
250
(Lethal
Dose)
100
100
100
100
100
100
0
100
100
100
100
0
Table 6-Id. Dicrotophos Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
tration Inhibition Average Deviation EC50
(mg/L) (%) (%) (%) (mg/L)
Standard
Inhibition Average Deviation EC50
(%) (%) (%) (mg/L)
1
13
I"4 5 7 5
9
160
0
4 2 3
0
6
200
11
14 ® 8 3
8
10
2 6 4
4
41
39
140 35 37 3
34
39
35
34 3
32
33
1-400 f
(Lethal ° 82 1
Dose) 81
80
TO 80 2
79
21
-------
Table 6-le. Thallium Sulfate Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
tration Inhibition Average Deviation EC50
(mg/L) (%) (%) (%) (mg/L)
Standard
Inhibition Average Deviation EC50
(%) (%) (%) (mg/L)
3
14
2.4 5 7 5
4
NA(a)
2
8 4 3
3
2
NA
3
24 I 4 2
6
3
2
5 6 3
9
6
240 8 6 1
4
16
18 17 1
16
17
27
2,400
(Lethal ° 17 7
Dose) j j
41
21 32 6
28
(a) NA = Not applicable.
Table 6-If. Botulinum Toxin Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
tration Inhibition Average Deviation EC50
(mg/L) (%) (%) (%) (mg/L)
Standard
Inhibition Average Deviation EC50
(%) (%) (%) (mg/L)
-6
0.00030 A -4 2
-2
NA(a)
-2
i -2 1
0
NA
3
0.0030 ~2 -1 3
-3
3
~-l -1 2
-2
-1
0.030 ^-2 3
-6
-4
7
;2 o 5
-1
4
0.30 ^
(Lethal 2 0 3
Dose) j
-2
-5
(a) NA = Not applicable.
22
-------
Table 6-lg. Ricin Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen-
Standard
Standard
tration
Inhibition Average
Deviation
EC-50
Inhibition Average
Deviation
ec50
(mg/L)
(%)
(%)
(%)
(mg/L)
(%)
(%)
(%)
(mg/L)
-2
-5
0.015
-1
2
1
0
2
-2
-1
0
-2
2
3
3
0.15
-3
2
0
3
-7
1
0
4
-1
NA(a)
1
NA
2
-1
1.5
-2
0
-1
-1
2
-4
-2
-7
-4
3
15
-1
-2
1
-2
0
(Lethal
Dose)
D
-1
2
0
2
-1
2
(a) NA = Not applicable.
Table 6-lh. Soman Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
tration Inhibition Average Deviation EC50
(mg/L) (%) (%) (%) (mg/L)
Standard
Inhibition Average Deviation
(%) (%) (%)
EC-50
(mg/L)
1
0.000068 "J -4 5
-9
NA(a)
-9
NA
5
0.00068 1 1 3
1
-4 0 4
-2
-2
0.0068 ^-2 2
0
-1
^-2 2
-3
-3
0.068(b) 3
(Lethal 7. -1 4
Dose) ^
2
3
-2 ° 3
-2
(a) NA = Not applicable.
-------
Table 6-li. VX Percent Inhibition Results
Results after 5 minutes
Results after 15 minutes
Concen-
Standard
Standard
tration
Inhibition Average
Deviation
EC-50
Inhibition
Average
Deviation
EC-50
(mg/L)
(%)
(%)
(%)
(mg/L)
(%)
(%)
(%)
(mg/L)
-1
4
0.00022
5
2
1
2
2
4
2
1
3
2
3
-2
0.0022
11
20
14
8
6
16
9
9
20
NA(a)
18
NA
-2
-3
0.022
0
1
12
3
7
-3
1
-1
-2
2
0.22
32
33
(Lethal
Dose)
-4
-2
0
7
17
-4
-1
6
18
(a) NA = Not applicable.
6.1.2 Potential Interferences
Table 6-2 presents the results from the samples that were analyzed to test the effect of potential
interferences on Microtox®. Aluminum and iron exhibited percent inhibitions near zero,
indicating little or no response to those compounds, while manganese exhibited inhibitions of
about 9% and zinc at 5 minutes had an inhibition at 6%. Exhibiting higher inhibitions were
copper, 40% and 61% at 5 and 15 minutes, respectively, and zinc, 28% at 15 minutes only.
All of the contaminant and potential interference samples were prepared in the DDW and
compared with unspiked DDW. Therefore, any background inhibition in the DDW was
mathematically corrected. To investigate whether Microtox® is sensitive to by-products of
disinfecting processes, DDW from water systems that use chlorination and chloramination were
analyzed and compared with ASTM Type IIDI water as the control sample. This determination
is crucial because the ability of Microtox® to detect toxicity is dependent on the light production
of the reagents in a clean drinking water matrix. If clean drinking water produces 100%
inhibition of light, the detection of subsequently added contaminants would not be possible. On
average, the chlorinated sample exhibited no detectable inhibition, indicating no toxicity, while
the chloraminated sample exhibited nearly complete inhibition (87% and 98% at 5 and
15 minutes, respectively. This suggests that samples that have been disinfected by using a
chloramination process are likely to interfere with the Microtox® results because the background
water sample would completely inhibit the Microtox® reagent.
24
-------
Table 6-2. Potential Interferences Results
Results after 5 minutes
Results after 15 minutes
Concen- Standard
Potential tration Inhibition Average Deviation
Interferences (mg/L) (%) (%) (%)
Standard
Inhibition Average Deviation
(%) (%) (%)
3
2
Aluminum 0.36 ^ 2 1
2
8
~3 1 5
0
0
37
Copper 0.65 ^ 40 3
38
62
62 ai 1
60
59
-2
Iron 0.069 2 ^
-1
-4
i -5 2
-4
8
Manganese 0.26 9 2
11
9
8 9 3
5
13
5
Zinc 3.5 ^ 6 1
6
29
g -
26
-11
Chlorination (a) -11
by-products 2
1
-15
- »
-10
Chloramin- ^
ation by- NA ^ 87 1
products QO
OO
O
OO
00 00 00 00
o\ o\ o\ G\
(a) NA = Not applicable.
6.1.3 Precision
Across all the contaminants and potential interferences, the standard deviation was measured
and reported for each set of four replicates to evaluate the Microtox® precision. The standard
deviation of the four replicate measurements was greater than 10% for only one sample and, in
most cases, it was less than 5%.
25
-------
6.2 Toxicity Threshold
Table 6-3 gives the toxicity thresholds, as defined in Section 5.2, for each contaminant. The
lowest toxicity threshold concentration was for cyanide at 0.25 mg/L after the 15-minute
reaction time, indicating that Microtox® was most sensitive to cyanide. For botulinum toxin,
ricin, soman, and VX, no inhibition greater than the negative control was detected, regardless of
the concentration level, indicating that the technology was not highly responsive to these
contaminants.
Table 6-3. Toxicity Thresholds
Contaminant
Concentration (mg/L)
(a)
Aldicarb
Colchicine
Cyanide
Dicrotophos
Thallium sulfate
Botulinum toxin
Ricin
Soman
VX
28
240
25 (5 minutes)
0.25 (15 minutes)
140
ND(b) (5 minutes)
240
ND
ND
ND
ND
(a) Unless otherwise noted, toxicity thresholds were the same at 5 and 15 minutes.
(b) ND = Significant inhibition was not detected.
6.3 False Positive/Negative Responses
False positive responses were observed for unspiked chloraminated tap water. As described in
Section 6.1.2, for a clean tap water sample that had been disinfected using a chloramination
process, Microtox® reported almost complete inhibition (87% and 98%). By-products of the
chloramination process apparently inhibited the Microtox® reagent. The water sample treated by
chlorination and then subsequently dechlorinated caused no detectable inhibition. Similarly, the
method blank samples caused no significant decrease of absolute light units and, therefore, were
not the cause for any false positive responses.
A false negative response is when a lethal dose of contaminant is present in the water sample
and no inhibition is detected. Table 6-4 gives each contaminant's lethal dose concentration and
shows whether or not each contaminant exhibited a false negative response at that concentration
level, Microtox® detected an inhibition induced by lethal doses of aldicarb, colchicine, cyanide,
dicrotophos, and thallium sulfate (15 minutes only), while botulinum toxin, ricin, soman, and
VX inhibitions were not detected at the lethal dose, indicating false negative responses.
26
-------
Table 6-4. False Negative Responses
Contaminant
Lethal Dose
Concentration
(mg/L)
False Negative
Response
Aldicarb
280
no
Colchicine
240
no
Cyanide
250
no
Dicrotophos
1,400
no
Thallium sulfate
2,400
yes (5 minutes)
no (15 minutes)
Botulinum toxin
0.30
yes
Ricin
15
yes
Soman
0.068(a)
yes
VX
0.22
yes
-------
Chapter 7
Performance Summary
Parameter
Compound
Lethal
Dose (LD)
Cone.
(mg/L)
Average Inhibitions at Concentrations
Relative to the LD Concentration
(% after 15 minutes) %
Range of
Standard
Deviations
(%)
Toxicity
Thresh.
(mg/L)'1"
LD
LD/10
LD/100
LD/1,000
Contaminants in
DDW
Aldicarb
280
81
31
4
3
2-5
28
Colchicine
240
12
2
5
3
1-3
240
Cyanide
250
100
96
46
8
0^1
0.25
Dicrotophos
1,400
80
34
6
2
2-4
140
Thallium
sulfate
2,400
32
17
6
4
1-6
240
Botulinum
toxin(b)
0.30
-4
0
-1
-2
1-5
ND(C)
Ricin(d)
15.0
-1
-4
0
-2
2-4
ND
Soman
0.068(e)
0
-2
0
-4
2-4
ND
VX
0.22
6
-2
9
3
2-18
ND
Potential
interferences in
DDW
Interference
Cone.
(mg/L)
Average Inhibitions at a
Single Concentration
(% after 15 minutes)
Standard
Deviation
(%)
Aluminum
0.36
1
5
Copper
0.65
61
1
Iron
0.069
-5
2
Manganese
0.26
9
3
Zinc
3.5
28
1
False positive
response
Nearly complete (87% and 98%) inhibition in water from a system disinfected by chloramination.
Inhibition due to contamination may not be detectable due to background inhibition. Inhibition
from water disinfected by chlorination was not significantly greater than zero.
False negative
response
No inhibition was detected for lethal doses of botulinum toxin, ricin, soman, and VX.
Other
performance
factors
The pictorial manual was useful, initial light measurements served as a good check of bacterial
health and instrument operation, sample handling was easy, and sample throughput was
15 samples per hour. Although the operators had scientific backgrounds, operators with little
technical training would probably be able to analyze samples successfully using only the
instructions as a guide. Microtox® was not tested in a non-laboratory setting because it is designed
to be only a laboratory benchtop instrument.
See Tables 6-la-i in the report for the precision around each individual inhibition result.
Lethal dose solution also contained 3 mg/L phosphate and 1 mg/L sodium chloride.
ND = Not detectable.
Lethal dose solution also contained 3 mg/L phosphate, 26 mg/L sodium chloride, and 2 mg/L sodium azide.
Due to the degradation of soman in water, the stock solution confirmation analysis confirmed that the concentration of the
lethal dose was 23% of the expected concentration of 0.30 mg/L.
28
-------
Chapter 8
References
1. Test/QA Plan for Verification of Rapid Toxicity Technologies, Battelle, Columbus, Ohio,
June 2003.
2. United States Environmental Protection Agency, National Secondary Drinking Water
Regulations: Guidance for Nuisance Chemicals, EPA/810/K-92/001, July 1992.
3. U.S. EPA Method 531.1, "Measurement of n-Methylcarbamoyloximes and
n-methylcarbamates in Water by Direct Aqueous Injection HPLC with Post Column
Derivatization," in Methods for the Determination of Organic Compounds in Drinking
Water—Supplement 111, EPA/600/R-95/131, 1995.
4. U.S. EPA Method 335.1, "Cyanides, Amenable to Chlorination," in Methods for the
Chemical Analysis of Water and Wastes, EPA/600/4-79/020, March 1983.
5. SW846 Method 8141 A, "Organophosphorous Compounds by Gas Chromatography:
Capillary Column Technique," Revision 1, September 1994.
6. U.S. EPA Method 200.8, "Determination of Trace Elements in Waters and Wastes by
Inductively-Coupled Plasma Mass Spectrometry," in Methods for the Determination of
Metals in Environmental Samples, Supplement I, EPA/600/R-94/111, 1994.
7. U.S. EPA Method 180.1, "Turbidity (Nephelometric)," Methods for the Determination of
Inorganic Substances in Environmental Samples, EPA/600/R-93/100, 1993.
8. American Public Health Association, et al. Standard Methods for the Examination of Water
and Wastewater. 19th Edition, 1997. Washington, DC.
9. U.S. EPA, Methods for Chemical Analysis of Water and Wastes, EPA/600/4-79/020.
10. U.S. EPA Method 524.2, "Purgeable Organic Compounds by Capillary Column GC/Mass
Spectrometry," Methods for the Determination of Organic Compounds in Drinking
Water—Supplement 111, EPA/600/R-95/131.
11. U.S. EPA Method 552.2, "Haloacetic Acids and Dalapon by Liquid-Liquid Extraction,
Derivatization and GC with Electron Capture Detector," Methods for the Determination of
Organic Compounds in Drinking Water—Supplement III EPA/600/R-95/131.
29
-------
12. Quality Management Plan (QMP)for the ETV Advanced Monitoring Systems Center,
Version 4.0, U.S. EPA Environmental Technology Verification Program, Battelle,
Columbus, Ohio, December 2002.
30
-------