Standard Operating Procedure for Use of the PetriSwiss
PS200 Instrument
SOP Number: EQ-13-00
Date Revised: 10-10-18

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 1 of 10
SOP Number
EQ-13-00
Title
Use of the PetriSwiss PS200 Instrument
Scope
This SOP describes the use of the semi-automated PetriSwiss PS200
to dispense agar based media into Petri dishes
Application
This SOP is used to dispense agar based media with the PS200
instrument


Approval Date
SOP Developer:

Print Name:
SOP Reviewer

Print Name:
Quality Assurance Unit

Print Name:
Branch Chief

Print Name:


Date SOP issued:

Controlled copy
number:

Date SOP withdrawn:


-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 2 of 10
TABLE OF CONTENTS
Contents	Page Number
1.
DEFINITIONS
3
2.
HEALTH AND SAFETY
3
3.
PERSONNEL QUALIFICATIONS AND TRAINING
3
4.
INSTRUMENT CALIBRATION
3
5.
SAMPLE HANDLING AND STORAGE
4
6.
QUALITY CONTROL
4
7.
INTERFERENCES
4
8. NON-CONFORMING DATA
4
9.
DATA MANAGEMENT
5
10.
CAUTIONS
5
11.
SPECIAL APPARATUS AND MATERIALS
5
12.
PROCEDURE AND ANALYSIS
5
13.
DATA ANALYSIS/CALCULATIONS
9
14.
FORMS AND DATA SHEETS
9
15.
REFERENCES
9

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 3 of 10
1. Definitions
1.	Screen = one of the several visual displays which may appear on the front
of the PS200.
2.	Key=is a display section on the screen. The most used keys are blue
squares with green symbols.
3.	Button= is a push button on the front of the PS200. There are two buttons:
one is labeled START/BRAKE and the other is labeled PUMP/MANU
[manual],
4.	Filling chamber= is the area under the hinged cover.
5.	Filling tubing assembly= consists of sections of plastic tubing with a metal
nozzle at one end, one open end which goes into the agar solution, and a
middle section of dual tubing for the two channels of the pump.
6.	Dish and plate =refer to the plastic Petri dish.
7.	Rack= is a removable tower which holds the plastic Petri dishes.
8.	The "carousel" or "carousel disk" is the carousel which holds the Petri
dish racks.
9.	The "dish separator" is the small carousel in the filling chamber.
2. Health and
Safety
Follow procedures specified in SOP MB-01, Laboratory Biosafety.
3. Personnel
Qualifications
and Training
Refer to SOP ADM-04, OPP Microbiology Laboratory Training.
4. Instrument
Calibration
1. The delivery volume may change after many autoclaving cycles because
of loss of tube flexibility, etc. The volume may be checked and adjusted:
a.	With the double tubing section of the filling tubing assembly
installed in the pump, put the nozzle end and the filling inlet end into
a container of water.
b.	Turn on power to get PS200/PS400 screen.
c.	Press green check-mark key to get the HOME screen.
d.	On the HOME page, press the PUMP key to get the PUMP screen.
e.	On the PUMP screen, press the key with the green arrow to fully
prime the pump.
f.	Then, press the CAL key to get the PUMP CALIBRATION screen.
g.	On the PUMP CALIBRATION screen, press the SPEED key until
the speed selection is green for the speed parameter of the method.

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 4 of 10

h.	Place the nozzle end of the filling tub to deliver into a graduated
cylinder.
i.	Then, press the key with the green arrow. The instrument will pump
the volume set in the method and display the PUMP CALIBRATION
screen with a "Volume Correction" option.
j. If the delivered volume is correct, press either of the two keys on the
far right. Then, press the key at the lower right to return to the
HOME screen.
k. If the delivered volume is not correct, enter the delivered volume on
the "Volume Correction" screen, and press green check key. The
instrument cycles to the PUMP screen.
1. On the PUMP screen the options include:
i.	The HOME key which causes the volume correction to be
saved.
ii.	The key with a green arrow which causes the corrected
volume to be dispensed so that the delivered volume may be
checked. Then, the HOME key may be pressed to save the
volume correction and return to the HOME screen,
iii.	CAL key which restarts the volume correction process. Note
that the previously entered volume correction is not saved.
5. Sample
Handling and
Storage
1.	Dispense agar into plates as specified by the individual preparation sheets.
2.	The temperature of the agar to be dispensed needs to be higher than for
manual pouring of plates because the agar will cool when travelling
through the filling tube and nozzle. Cooling will be greater between rack
changes and agar solution container changes when the agar is not flowing.
6. Quality Control
1.	The delivered volume should be calibrated when a new filling tube is used
and after a tube has been sterilized several times (approximately 15-20
times).
2.	The delivered volume should be checked by the user quarterly.
3.	Calibration is specific to the type/brand of Petri dish, conduct calibrations
accordingly.
7. Interferences
1.	Broken plates and plates loaded incorrectly.
2.	A change in silicone tube flexibility.
8. Non-
conforming
1. Management of non-conforming data will be consistent with SOP ADM-
07, Non-Conformance Reports.

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 5 of 10
Data

9. Data
Management
1. Data will be archived consistent with SOP ADM-03, Records and
Archives.
10. Cautions
1.	Do not put hands or fingers near the carousel, racks, or filling chamber
when an operation is being performed or initiated.
2.	Do not use harsh chemicals in cleaning. Agar spills may be removed with
a water-moistened towel or tissue. Disinfection may be done with a tissue,
towel or swab that is moistened with 70% ethanol in water.
3.	The agar will cool while travelling through the filling tubing. If the agar is
too cool, bubbles will appear in the plated agar.
4.	Do not use PetriSwiss instrument for dispensing agar media that requires
addition of enrichments, a heat-sensitive step e.g. 7H11 agar.
11. Special
Apparatus and
Materials
1.	Water bath to hold agar at 57±2° C. Monitor temperature of water bath by
using a thermometer inside a flask of water, kept inside the water bath.
2.	Use ring weights to stabilize the agar containers in the water bath.
3.	Polystyrene Petri dishes, size 100 mm x 20 mm, slippable, Excel
Scientific D-905, Sigma-Aldrich number P5606-400, or other comparable,
slippable Petri dishes.

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 6 of 10
12. Procedure and
Analysis
1.	The instrument should be operated according to manufacturer's
instructions.
2.	Instrument parameters for filling plates are contained in a software
program. Software programs may be created and edited (as noted below)
by users. Parameters in a program include the following:
a.	Filling volume: set according to media preparation sheet; typically,
25, 30 or 50 mL
b.	Dish height: set for 21.5 mm for polystyrene Petri dishes, size 100
mm x 20 mm, slippable, Excel Scientific D-905, Sigma-Aldrich
number P5606-400.
c.	Customary settings for the other parameters are:
i.	Pause Time: 0.0 seconds
ii.	Filling speed: Depends on tube size and calibration; typical
values are 663 mL/minute and 708 mL/minute.
iii.	Cooling Tower: not applicable (N/A)
iv.	UV Light Anti Drop: On
v.	Ink-Jet: N/A
vi.	Speed/Mix: Fast
12.1 Preparing for
the filling
procedure
a. Before using the PS200 for filling plates, disinfect (using a tissue,
towel or swab moistened with 70% ethanol) the dish separator, the
agar tray (the platform on which the dish separator rests), and the
brackets (inside and outside) that hold the filling nozzle in place.
Include the areas on which the pistons (dish lifters) rest. Use the
Manual keys to manipulate the unit as desired. The Manual keys are
available from the HOME screen (see below) by pressing the
MANUAL key. Close the hinged cover over the filling chamber.
12.2 Performing the
filling
procedure
a.	Turn on power to get PS200/PS400 screen.
b.	Press green check-mark key to get the HOME screen.
c.	On the HOME screen, press FILL key. Unit moves to position A and
displays FILL screen.
d.	On the FILL screen, press the green arrow key to begin operation
with the program shown. Get Set Dish Count screen/or
Press the LIST key to select a program/or
Press the EDIT key to make a change in the program or to check a

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 7 of 10
program parameter value (e.g., fill volume, etc.).}
e.	On the Set Dish Count screen, input number of dishes to fill (or a
number that is greater than the number anticipated). Press the green
arrow check-mark key to get the PREPARE FILLING screen.
(If a WARNING screen with "RESET DISH COUNT?" message
comes up, press the green check-mark key.)
f.	While on the PREPARE FILLING screen,
i.	Load racks of plates (Petri dishes) on their carousel, using the
left and right arrow keys at the center of the screen to move
the carousel for convenience in loading. (The first rack
should be empty to receive the plates to be filled.) The plates
may be loaded earlier, if desired.
ii.	Set the empty column rack to position A using the same keys.
iii.	Transfer the agar from the autoclave to the water bath, held at
57±2°C, if this has not been done earlier. Use weights to
stabilize the agar containers in the water bath.
iv.	Press the green arrow at the far right of the screen to get the
Process Status screen; the process begins.
g.	If it is desired to test the movement of plates without filling, do not
press the BRAKE button on the next step. Movement of plates will
stop when the plate count value is reached and the instrument will
return to the HOME screen.
h.	As the process begins, press the BRAKE button. The process begins
with moving of racks, checking the plate separator to make sure there
are no plates already there, and lowering the first plate to be filled.
The BRAKE key must be pushed before the first plate is in position
to be filled. When this process is completed, the PAUSE screen
appears.
i.	While on the PAUSE screen,
(Note: Steps i through iv may be performed earlier, if desired.)
i.	Move the lever above the pump to its right-most position,
ii.	Install the double tubing section of the (previously
autoclaved) agar filling tubing assembly into the two channels
of the pump, placing the nozzle end of the assembly on the
left,
iii.	Move the lever above the pump to its left-most position,

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 8 of 10
iv.	Remove the foil from the nozzle end of the assembly and
install the nozzle into its brackets. (There are two brackets for
the filling tube nozzle: one is on the perimeter of the hinged
cover over the pump and the second is inside the filling
chamber between the first bracket and the plate separator. The
nozzle gives a click sound when it is properly inserted in the
second bracket.)
v.	Remove the foil from the other end of the assembly and put
the tubing in the agar solution (which is in a water bath).
vi.	Press the PUMP button to fill the tubing (up to the nozzle)
with agar.
vii.	Press the green arrow key to get the FILLING PROCESS
screen. Note: if the filling is to be from more than one
container of agar, the first container may be moved out of the
water bath after the filling operation begins to prevent
contamination of the bath when the filling tube is moved to
the next container.
j. If an agar container has insufficient agar to fill the next plate, press
the BRAKE button to pause the filling process.
The display will show the PAUSE screen.
Remove the next agar container from the water bath and place the
filling tube in it.
Press the green arrow key to continue the filling operation on the
FILLING PROCESS screen.
Return the emptied container to the water bath to keep the remaining
agar from solidifying before the container is rinsed. Weights should
remain on the container to stabilize it in the bath.
k. If the number of plates filled reaches the dish count value entered
earlier, the filling operation stops but all the plates in the current rack
will be transferred.
1. If the agar supply is exhausted, press the orange key at the bottom
right of the FILLING PROCESS screen. The Process Status icon
will turn orange.
The unit will stop filling plates, but it will empty the current rack
before stopping.
m. When the unit has stopped filling plates, transfer the nozzle to a
container (e.g., the last agar container just used) and the tubing end to

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 9 of 10
q-
r.
s.
a container of (preferably hot) distilled water. This may be a 500-mL
container from the water bath.
When the unit has emptied the last rack and returned to the HOME
screen, press and hold the PUMP button to rinse the filling tubing
with about 500 mL of water.
Remove the filling tubing assembly from the pumping mechanism,
rinse it with distilled water, and set it aside for autoclaving.
Clean the filling area of the PS200:
i. Raise the hinged cover.
Remove the plate separator and the agar tray (the platform on
which the plate separator rests.
Clean away any spilled agar with a water-moistened towel.
Clean and disinfect with a tissue, towel or swab moistened
with 70% ethanol, the plate separator, the agar tray, the
pistons (dish lifters) and the brackets that hold the filling
nozzle in plate. Use the Manual keys to manipulate the unit
as desired.
Close the hinged cover.
Turn off power switch of the PS200.
Clean the emptied agar solution container(s).
Allow the dishes to remain in their racks until the agar has solidified.
li.
in.
IV.
v.
12.3 Sterilization of
filling tubing
assembly
a.
b.
c.
d.
e.
Dry the filling tubing assembly before autoclaving. (Compressed air
may be blown through the tubing assembly to dry it.)
Wrap aluminum foil around the nozzle and at least six inches of the
tubing where the nozzle is attached.
Wrap aluminum foil around at least 14 inches of the other end of the
tubing assembly.
Seal the assembly in an autoclave pouch.
Autoclave the assembly at 121° C for either 20 or 25 minutes on a
gravity cycle. (Time is selected based on any other items that may be
conveniently autoclaved at the same time.)
13. Data Analysis/
Calculations
None
14. Forms and Data
Test sheets are stored separately from the SOP under the following file name:

-------
SOP No. EQ-13-00
Date Revised 10-10-18
Page 10 of 10
Sheets
PetriSwiss Tubing Calibration Log EQ-13-00_Fl.docx.
15. References
1. PetriSwiss PS-200 User Manual 1,0-E, Version 2.0

-------