United States	Office of Underground Storage
Environmental Protection Tanks
oERft
Standard Test Procedures For
Evaluating Release Detection
Methods: Volumetric And Non-
volumetric Tank Tightness
Testing
May 2019

-------
Acknowledgments
The U.S. Environmental Protection Agency's Office of Underground Storage Tanks contracted
with Battelle under Contract No. EP-C-10-001 to revise EPA's 1990 Standard Test Procedures
for Evaluating Various Leak Detection Methods. Individual members of the National Work
Group on Leak Detection Evaluations, as well as Ken Wilcox and Associates, reviewed this
document and provided technical assistance. A stakeholder committee, comprised of
approximately 50 representatives from release detection method manufacturers and various
industry associations, also commented on this document.
li

-------
Contents
Acknowledgments	ii
List Of Acronyms And Abbreviations	vi
Section 1: Introduction	1
1.1	Background	1
1.2	Objectives And Applications	1
1.3	Evaluation Approach Summary	2
1.3.1	Volumetric TTT Methods	2
1.3.2	Non-Volumetric TTT Methods	3
1.3.3	Sensors	3
1.3.4	Effects Of High Groundwater Level And Considerations For
Double Walled Tanks	4
1.4	Organization Of This Document	5
Section 2: Safety	6
Section 3: Apparatus And Materials	7
3.1	Tank Tightness Test Method Equipment	7
3.2	Tanks	7
3.2.1 Tank-Related And Other UST System Components	8
3.3	Product	9
3.4	Leak Simulation Equipment	11
3.4.1	Leak Simulation Approach For Non-Volumetric Methods To
Include Acoustical And Pressure-Vacuum Decay Methods	11
3.4.2	Leak Simulation Approach For Tracer Methods	12
3.4.3	Leak Simulation Approach For Tightness Test Methods Using
An Optical Device	13
3.5	Sensor Evaluation Equipment	14
3.5.1	Liquid Sensor Test Vessel	14
3.5.2	Vapor And Pressure Decay Test Vessel And Leak Simulation	14
3.6	Miscellaneous Equipment	15
Section 4: Test Procedures	16
4.1	Environmental Data Records	17
4.2	TTT Evaluation Test Procedures	18
4.2.1	Volumetric TTT Methods	18
4.2.2	Non-Volumetric TTT Methods	21
4.3	Implementation Of The Test Procedures	24
4.3.1	Application Of The Test Procedure To Acoustical Methods	26
4.3.2	Application Of The Test Procedure To Tracer Methods	29
4.4	Testing Problems And Solutions	31
4.5	Sensor Evaluation Test Procedures	32
4.5.1	Liquid Phase Sensor Test Procedures	33
4.5.2	Product Vapor Phase Sensor Test Procedures	37
in

-------
4.5.3 Test Procedures For Tightness Testing Using A Vacuum
Monitor On A Double-Walled Tank Interstice With Or Without
The Addition Of A Liquid Sensor	38
4.5.4	Recovery Time	40
4.5.5	Test Procedures For Tightness Testing On A Liquid Filled
Interstice Of A Double-Walled Pipeline Using A High Pressure
And A Low Pressure Limit Switch Sensor	40
Section 5: Calculations	46
5.1	Estimation Of The Volumetric Method Performance Parameters	46
5.1.1	Basic Statistics	46
5.1.2	False Alarm Rate, P(fa)	48
5.1.3	Probability Of Detecting A Leak Rate Of 0.10 gal/hr, P(d)	50
5.2	Estimation Of The Non-Volumetric Method Performance Parameters	50
5.2.1	False Alarm Rate, P(fa)	50
5.2.2	Probability Of Detecting A Leak, P(d)	51
5.3	Other Reported Calculations	52
5.4	Supplemental Data Analyses (Optional)	54
5.5	Sensor Performance Calculations	55
Section 6: Interpretation	58
6.1	Basic Performance Estimates	58
6.2	Limitations	58
6.3	Additional Calculations	59
Section 7: Reporting Of Results	60
Appendices
Appendix A: Definitions And Student's t Distribution	A-l
Appendix B: Volumetric Methods Reporting Forms	B-l
Appendix C: Non-Volumetric Methods Reporting Forms	C-l
Appendix D: Sensor Evaluation Forms	D-l
Figures
Figure 1. Example Schematic Of A Vapor Test Chamber	15
Figure 2. Student's t-Distribution Function	49
IV

-------
Tables
Table 1. Analytical Methods For Bio-Component Determination	11
Table 2. Leak Rates To Evaluate A Method At 0.10 gal/hr Leak Rate	18
Table 3. Leak Rate And Temperature Differential Volumetric Test Design	20
Table 4. Leak Rate And Temperature Differential Non-Volumetric Test Design	23
Table 5. Notation Summary	47
Table 6. One Sided Confidence Limits For P(fa) And P(d)	54
Table 7. Performance Parameters	55
Table 8. Notation Summary For Water Sensor Readings At The jth Replicate	56
v

-------
List Of Acronyms And Abbreviations
ASTM International American Society for Testing and Materials (ASTM) International
ATGS
automatic tank gauging system
B
bias
°C
degree Celsius
CFR
Code of Federal Regulations
cm
centimeter
df
degrees of freedom
°F
degree Fahrenheit
gal/hr
gallon per hour
LR
leak rate
mL/min
milliliter per minute
MSE
mean square error
P(d)
probability of detecting a leak
P(fa)
probability of false alarm
ppmv
parts per million by volume
psi
pounds per square inch
SD
standard deviation
T
temperature differential
Th
threshold
TTT
tank tightness testing
UST
underground storage tank
VI

-------
Section 1: Introduction
1.1	Background
The federal underground storage tank (UST) regulation in 40 Code of Federal Regulations (CFR)
Part 280 specifies performance standards for release detection methods. UST owners and
operators must demonstrate that the release detection method they use meets the U.S.
Environmental Protection Agency's (EPA) regulatory performance standards. This document
provides test procedures for evaluating tank tightness testing (TTT) methods.
This tank tightness testing document is one of four EPA standard test procedures for release
detection methods. The test procedures present performance testing approaches to evaluate
various release detection method categories against the federal UST regulation in 40 CFR Part
280, Subpart D. To provide context for the four test procedure documents, EPA developed
General Guidance For Using EPA 's Standard Test Procedures For Evaluating Release
Detection Methods. The general guidance provides an overview of the federal UST regulation,
methods, and testing that may result in release detection methods listed as compliant with the
regulatory performance standards. The general guidance is integral; it must be used with the test
procedures.
Tank tightness testing methods must be capable of detecting a leak of 0.10 gallon per hour
(gal/hr) with a probability of at least 95 percent, while operating at a false alarm rate of 5 percent
or less. There are two categories of TTT methods:
•	Volumetric testing methods, which quantify the leak rate in gal/hr, and
•	Non-volumetric testing methods, which report the qualitative assessment of leaking or
not leaking against a threshold.
These two categories require different testing and statistical analysis procedures to evaluate their
performance. This document also presents testing of sensors as components of release detection
systems for their performance, such as sensitivity and specificity. You may use this document to
evaluate methods other than tank tightness test methods. Other certified leak rates may be used.
For example, manufacturers may opt to evaluate leak rates other than those established in the
federal UST regulation. In those scenarios, adjust the evaluation accordingly. The evaluator
ensures reporting forms and other relevant documents are modified, as required, to indicate and
provide appropriate details relevant to the UST system component evaluated.
1.2	Objectives And Applications
This test procedure addresses two objectives. It provides procedures to test TTT methods in a
consistent and rigorous manner. Also, it allows the regulated community and regulatory
authorities to verify compliance with the federal UST regulation. Tank owners and operators
must demonstrate that the method of release detection they use meets EPA's performance
standards.
1

-------
This procedure evaluates methods that test tanks or tanks connected by siphon piping at a
specific point in time. The procedure considers a number of factors encountered at UST
facilities such as temperature and leak rate. Volumetric methods quantify leak rates to determine
if a tank is leaking. Non-volumetric methods determine a yes or no answer to the question: Is
the tank leaking? Commercially available non-volumetric methods rely on one or more physical
results from a leaking tank to make this determination. These include acoustical, optical, tracer,
and pressure decay methods. Some TTT methods use various sensors to detect the presence of
liquid, vapor, or change in liquid level. We discuss the majority of sensor test procedures in
Section 4.5; however, water ingress testing is in EPA's Standard Test Procedures For
Evaluating Release Detection Methods: Automatic Tank Gauging Systems.
You can find information on sensor types and other general information regarding sensors in
General Guidance For Using EPA 's Standard Test Procedures For Evaluating Release
Detection Methods.
Although safety is a consideration while conducting testing, these test procedures do not address
the issue of safety specific to detection methods and their operating procedures, merely basic
laboratory safety concerns and procedures. The vendor is responsible for conducting the testing
necessary to ensure that method equipment is safe for operation and capable of being used with
the intended product.
Ultimately, you can use the results from this procedure to prove that the method meets the
requirements of 40 CFR Part 280 and is subject to limitations listed on EPA's standard
evaluation form in Appendix B or C for volumetric and non-volumetric, respectively.
1.3 Evaluation Approach Summary
1.3.1 Volumetric TTT Methods
Set up a volumetric TTT method in the test tank to measure a leak rate under a no-leak or tight
condition with three induced leak rates of 0.05, 0.10, and 0.20 gal/hr. You must conduct a
minimum of 24 tests. You must partially empty the tank to half full or less, and then refill it to
the 90-95 percent full level for at least every other test. When filling the tank to the test level,
use product at these three different temperatures:
•	At least 10 degrees Fahrenheit (°F) warmer than that in the tank for one third of the
filling;
•	At least 10°F cooler than that in the test tank for one third of the filling; and
•	At the same temperature as the in-tank product for one third of the filling.
The volumetric test method's ability to track actual volume change is determined by the
difference between the volume change rate measured by the test method and the actual induced
volume change rate for each test. From these differences, you can calculate the probability of
false alarm (P(fa)) and the probability of detecting a leak (P(d)). Report performance results on
the Results Of U.S. EPA Standard Evaluation Volumetric Tank Tightness Testing Method form in
Appendix B.
2

-------
1.3.2 Non-Volumetric TTT Methods
Set up and use a non-volumetric TTT method test tank following the method's standard
operating procedure. Conduct a minimum of 21 independent tests of the tank under the no-leak
condition. Use the results of these tight tank tests to estimate the P(fa). In addition, induced
leaks at rates not to exceed 0.10 gal/hr are simulated. Conduct a minimum of 21 independent
tests with induced leaks. Keep the simulation condition of tight tank or induced leaks blind to
the vendor and randomized in the test design. Compare this reported result with the actual
condition of the tank during testing to estimate the P(fa) and P(d). Report performance results on
the Results Of U.S. EPA Standard Evaluation Non-Volumetric Tank Tightness Testing Method
form in Appendix C.
The method accurately detects a leak of the specified size, for example 0.10 gal/hr, in the
presence of interference. Do not include sources of interference, such as product temperature
changes that do not affect the operation of a method, in the testing. However, the evaluator must
consider other sources of interference, such as vibrations from traffic, that may affect the
operation of the method and include tests to determine whether the method can successfully
overcome these sources of interference. These tests are designed to cover interference conditions
encountered in approximately 75 percent of real-world tests. You do not need to include sources
of interference that are rarely encountered in the field.
Some non-volumetric test methods use more than one approach to detect a leak. For example,
some vacuum-based methods use a capacitance sensor to check for water ingress that would
indicate a leak in the tank as well as check for an acoustic signal. Test and evaluate each
approach to determine whether or under what conditions the method meets EPA's performance
standards.
1.3.3 Sensors
These test procedures provide multiple test designs to evaluate release detection sensor
capabilities. Depending on the equipment, its intended use by the vendor, and input from the
evaluator, the appropriate test designs will provide data on the specificity and sensitivity of the
sensor. In general, a sensor reacts to a change in the environment in which it is located. Many
sensors do not come in contact with product and are not expected to perform differently with
various products stored in an UST system. Many sensors are non-discriminating, in that they
react to the change whether it is in contact with water or product. However, in cases where a
sensor is designed to react to a change in electric potential, such as capacitance and conductivity
sensors, in a system storing alcohol blends, the sensor may not function at a specific set point
due to interference by water. Also, in cases where a sensor is designed to react to the presence of
hydrocarbons and comes in contact with liquid or vapor product in a system storing ethanol
blends, the sensor may not function if the ethanol component of fuel is high enough to dilute the
hydrocarbon component. Furthermore, in high alcohol blends such as E85, ethanol could absorb
enough water in a system that has an abundance of water where the sensor might not function as
intended. The sensor might indicate water instead of fuel and an alarm condition associated with
the presence of fuel could exist that could be missed by the sensor.
3

-------
These test procedures evaluate methods that provide liquid level measurement either from a wet
hydrostatic environment such as a brine-filled interstice, or a dry environment such as a sump
pit or secondary containment under vacuum or ambient conditions. Sensor test procedures must
evaluate the sensor's ability to identify that a liquid is present, detect a change in liquid, or
identify either water or hydrocarbons specifically. Additionally, discriminating sensors are
sensors monitoring wet or dry spaces that may detect product or product vapor.
Depending on the capabilities of the sensor, test liquid sensors by introducing liquid such as
water or product into the dry vessel or into a vessel containing water or product. When product
is on top of the water, determine the detection limit of the sensor by the thickness of the product
layer. When you add water to product, test the sensors to detect water entrained in the product or
in a separated phase on the bottom. In addition to the procedures provided in Section 4.5, the
evaluator may use the test procedures for detecting water entering product as a separate phase or
entrained in the fuel as presented in EPA's Standard Test Procedures For Evaluating Release
Detection Methods: Automatic Tank Gauging Systems. Testing procedures for sensor
functionality in systems with alcohol blends must include testing with a variety of amounts of
water to determine whether water interferes with performance of sensors designed to react to a
change in electric potential, such as for capacitance and conductivity sensors. At minimum, the
evaluator must test the vendor's desired alcohol blend and that alcohol blend with three water
mixtures: 80 percent alcohol blend and 20 percent water; 60 percent alcohol blend and 40
percent water; and 30 percent alcohol blend and 70 percent water. For sensors that discriminate
between hydrocarbons and water intended to be used with alcohol blends, the evaluator must
evaluate the discriminatory sensor both with alcohol blend fuel that is fully in solution with
water, as well as with distinct phase separation layer with neat gasoline on top. The sensor may
only detect a certain layer or layers.
Test vapor sensors in a more controlled and contained test chamber using various concentrations
of hydrocarbon mixtures or hydrocarbon-alcohol blends for sensors used in alcohol blends to
determine the performance parameters of the sensor. Where sensors are designed to detect the
presence of liquid hydrocarbons intended to be used for alcohol blends, repeat the test
procedures presented in Section 4.5 with a variety of blends as determined by the evaluator, with
input from the vendor, to determine the accuracy and specificity of sensors range of operability
in alcohol blends. At minimum, the evaluator must test the vendor's alcohol blend the sensor is
intended to be used with and that alcohol blend with three water mixtures: 80 percent alcohol
blend and 20 percent water; 60 percent alcohol blend and 40 percent water; and 30 percent
alcohol blend and 70 percent water.
1.3.4 Effects Of High Groundwater Level And Considerations For Double Walled Tanks
The groundwater level is a potentially important variable in tank tightness testing. Groundwater
levels may be above the bottom of the tank, particularly in coastal regions where tidal effects
may cause fluctuations in the groundwater level during testing. If the groundwater level is above
the bottom of the tank, the water pressure on the exterior of the tank tends to counteract the
product pressure from inside the tank. If the tank has a leak or hole below the groundwater level,
the leak rate in the presence of a high groundwater level will be less than with a lower
4

-------
groundwater level. If the groundwater level is high enough, water may intrude into the tank
through the hole.
The evaluator must document how the method addresses the groundwater level or in the case of
testing the primary containment of a double walled tank, how the presence of a liquid filled or
closed interstice may affect the method. A method that does not account for groundwater level
or interstitial issues for a double-walled tank is inadequate. If the groundwater or brine level is
above the bottom of the tank, a testing condition must accommodate for the high groundwater or
brine level. The evaluator can do this by ensuring the tank has an outward pressure throughout
or that groundwater or brine exerts an inward pressure at all levels in the tank. If the method
uses an alternative approach to compensate for groundwater or brine levels, the evaluator must
perform an engineering evaluation of the approach to ensure it is adequate. If testing the primary
containment of a double-walled tank with a vacuum or pressure method, the evaluator or method
must assure that the interstice is open whether it is dry or not. If in doubt, the evaluator may
require additional tests to those detailed in this document.
1.4 Organization Of This Document
This document is organized as follows:
•	Section 2 presents a brief discussion of safety issues.
•	Section 3 discusses the apparatus and materials needed for the evaluation of the test
methods.
•	Section 4 presents the step-by-step procedures for volumetric and non-volumetric test
methods and sensors.
•	Section 5 describes the data analysis.
•	Section 6 provides some interpretation of the results.
•	Section 7 describes how to report the results.
Four appendices are included in this document:
•	Appendix A includes definitions of some technical terms.
•	Appendix B presents forms for volumetric methods.
•	Appendix C contains similar forms for non-volumetric methods.
•	Appendix D contains forms for sensor testing and reporting.
The forms in Appendices B and C form the basis of the standard evaluation report including: a
standard reporting form for the evaluation results, a standard form for describing the operation of
the method, data reporting forms, and an individual test log.
5

-------
Section 2: Safety
The vendor tests the TTT equipment and ensures it is safe for its intended use. As part of a
standard operating procedure, the vendor provides a safety protocol for each release detection
method. The protocol specifies requirements for safe installation and use of the method. In
addition, all facilities hosting an evaluation provide a safety policy and procedure to the
evaluator and staff on site. You must follow all safety requirements to ensure the safety of those
performing the evaluation and those near the evaluation-testing site.
At a minimum, ensure this safety equipment is available at the site:
•	Two class ABC fire extinguishers;
•	One portable eyewash station;
•	Adequate quantity of spill absorbent; and
•	Appropriate safety signs such as No Smoking, Authorized Personnel Only, and Keep Out.
Follow all safety procedures appropriate for the product in the tanks and test equipment.
Personnel working at the UST facility must wear safety glasses when working with product and
steel-toed shoes when handling heavy pipes or covers. Place the safety equipment at the site;
before work begins, post the No Smoking, Authorized Personnel Only, and Keep Out signs.
These test procedures only address the issue of the method's ability to detect leaks. They do not
address testing the release detection method for safety hazards. The vendor is responsible for
meeting other construction standards testing that addresses key safety hazards such as fire,
shock, intrinsic safety, and product compatibility.
6

-------
Section 3: Apparatus And Materials
3.1	Tank Tightness Test Method Equipment
The vendor supplies equipment for each TTT method tested. In general, the equipment includes
a means of monitoring the tank using vacuum or pressure decay, acoustical methods, or detection
of water to indicate a leak and instrumentation for collecting and recording the data to interpret
the result as a pass or fail for the tank. For tracer methods, equipment includes a means for
introducing the tracers into the tank or the backfill.
Trained personnel who regularly use the method and are deemed qualified by the vendor to
perform commercial tests should conduct the test. This ensures the vendor's method is operated
properly and eliminates problems that newly trained or untrained individuals may have with the
equipment. If the equipment owner normally operates the method, then the equipment owner
provides personnel to operate the method. If applicable, follow the vendor's standard quality
control methods when performing the tests.
3.2	Tanks
The evaluation test procedures require that the UST is tight. You will need a second tank or a
tank truck to store product for the emptying and refilling cycles. The tank must be tested and
proved tight by another release detection method. The tank must not have a history of problems.
In addition, the test procedures call for an initial trial run with the test method under stable
conditions. Before testing begins, the trial run is used to confirm the tank tests tight; if it does
not, there may be a problem with the tank or the test method, which must be resolved before
proceeding with the evaluation.
The tank facility used for testing must have at least one monitoring or observation well. The
primary reason is to determine the groundwater level. The presence of groundwater above the
bottom of the tank will affect the leak rate in a real leak situation, that is, the flow of product
through a hole in the tank wall. The flow is a function of the differential pressure between the
inside and outside of the tank. It is not necessary to require that testing against the evaluation
test procedure occur in a tank entirely above the groundwater level; however, it is important to
record the groundwater level if an actual leak occurs during testing.
Volumetric methods that measure volume or level changes of liquid product occurring as a result
of a leak generally perform worse as the size of the tank increases. The evaluation may use tanks
of any size. The results of the evaluation are applicable to all smaller tanks; therefore, the larger
the test tank, the broader the applicability of the evaluation. For the majority of methods, the
results also apply to larger tanks, but are restricted to tanks no more than 50 percent larger in
capacity than the test tank for single tanks and 25 percent larger for tanks connected by siphon
piping. However, the accuracy of some test methods and test method categories are very volume
sensitive while others are much less sensitive to volume. For that reason, it is appropriate to
impose correspondingly more or less restrictive tank size applicability for certain methods. For
example, upscaling results from test tank sizes used for vacuum decay-based methods are
7

-------
generally not appropriate. The evaluator must provide justification if upscaling is applied.
Conversely, test tank sizes used for tracer-based methods presumably have no limit, provided
that the dose applied is appropriately measured. Therefore, there should not be a limit to
upscaling results. The evaluator must provide an explanation for applicability or non-
applicability of upscaling results to larger tanks in the reporting form.
For both volumetric and non-volumetric methods, the evaluator determines the appropriate size
limit based on subsequent testing, physical principles involved, and other available data and
states the limit on the results forms. For example, tanks larger than 50,000 gallons may have a
different construction and geometry than standard horizontal cylindrical tanks. The tank
geometry and construction may impose limits rather than the size.
For tracer methods, the characteristics of a tank are less important. However, the test tank used
for the evaluation must be tight. The primary purpose of the test tank is to provide an
environment, which is representative of typical tank installations. The test tank is important for
testing for false alarms. The procedure of adding and mixing tracer to the product is a potential
source of false alarms from inadvertent release of tracer into the environment.
3.2.1 Tank-Related And Other UST System Components
It is also possible to perform a VTTT or NVTTT evaluation on tank-related and other UST
system components such as a tank interstice, pipeline interstice, or containment sump. The
technologies available include, but are not limited to liquid sensors, pressure sensors, vacuum
sensors, level measurement devices, and optical devices. Many of the requirements for
evaluating a tank tightness test method for tank-related and other UST system components are
the same as when evaluating a method that tests the primary space of a tank. The evaluator
determines whether thermal conditioning is needed to adequately evaluate a specific tightness
test method and provides the rationale for the determination. For example, some leak detection
methods are used primarily to perform a tightness test on a liquid filled interstice of a newly
installed double-walled tank where no product is present in the primary space of the tank. It may
be unnecessary to implement thermal conditioning for such a system, assuming that the tank has
been installed in the ground for at least 24 hours. In addition to newly installed tanks, it may be
unnecessary to implement thermal conditioning on a tightness test method that requires the tank
to be stable without any fuel deliveries for 24 hours.
If thermal conditioning is necessary, then follow the test schedule on Table 3 for volumetric
methods and Table 4 for non-volumetric methods. If thermal conditioning is required in addition
to the test schedule to implement thermals, also consider the volume of test apparatus since that
is a factor when measuring the level in the reservoir of the interstice. Consider the volume of the
test vessel used in the evaluation where thermal conditioning is a factor and when placing a limit
on the volume the tightness test method is applicable to when testing a tank interstice. Interstitial
volumes smaller than the evaluation test vessel are acceptable. For larger interstitial volumes,
limit the tightness test method to 50 percent larger than the test vessel used in the evaluation.
When performing an evaluation of a tightness test method for a liquid filled interstice, a test
vessel with a liquid filled reservoir connected to the top of the vessel may be used in a lab
8

-------
environment. If the evaluator verifies that thermal conditioning is unnecessary and other
potential sensitive parameters are determined to be negligible, then there is no volume limit
placed on the test method when testing a tank interstice; that means, a test vessel of
approximately 5 gallons or more may be used. However, if the evaluator determines that thermal
conditioning is necessary or other potential sensitive parameters cannot be ignored, the volume
of the test vessel must be carefully measured since there will be a limit placed on the volume that
the test method is applicable to when testing a tank interstice. In this case, the test vessel can be
of any volume ranging from 5 gallons up to 1,000 gallons or more, comparable to the interstitial
space of actual tanks. The evaluator must explain on the reporting form about applicability or
non-applicability of upscaling results to larger tanks.
When performing an evaluation of a tightness test method for a liquid filled interstice, you may
use a test vessel of a known volume in a lab environment where temperatures can be constantly
controlled. The test vessel can be any volume ranging from 5 gallons to 1,000 gallons with a
liquid filled reservoir connected to the top of the test vessel. The size and surface area of the
reservoir should be a size that is commonly used with a liquid filled double-walled tank
interstice. Measure the size of the reservoir carefully in order to calculate the surface area.
Typically, the larger the surface area of the reservoir, the longer the test duration will be, so you
must report the surface area. You may also calculate the test time for some methods on
reservoirs with a different size surface area.
When performing an evaluation on a tightness test method for containment sumps, you may
perform the evaluation in a lab environment in a test vessel that is similar to the size and
dimensions of a typical containment sump. It may be unnecessary for thermal conditioning
when performing an evaluation on a containment sump. Perform a total of 42 tests on a non-
volumetric test method including 21 tight condition tests and 21 with a leak present that is
calibrated to 0.10 gal/hr or less. The federal UST regulation does not establish leak rates for
liquid tight containment sump testing. You may evaluate for target leak rates other than 0.10
gal/hr. For a volumetric tightness test method, perform 24 tests including 6 tests in the tight
condition and 18 tests with a leak present. Of the 18 tests, perform 6 at each of the required
leaks rates of 0.05, 0.10, and 0.20 gal/hr for 0.10 and 0.20 gal/hr listed leak detection rates. The
evaluator will choose appropriate leak rates for a target listed leak rate other than 0.10 or 0.20
gal/hr.
3.3 Product
The most common products in USTs today are motor fuels, particularly non-alcohol blended
gasoline, alcohol-blended gasoline, diesel, and biodiesel fuels. These test procedures are
designed to evaluate currently widely marketed products using at least 24 tests for volumetric
methods and at least 42 tests for non-volumetric methods.
The evaluator and the vendor choose the product, but it must be capable of being used with the
release detection method. Testing a method with a specific product verifies its performance with
that product. However, you may use products with similar physical and chemical characteristics,
but you must use caution when inferring that results represent typical responses across products.
9

-------
The evaluator must justify the extent of applicability of results to other products. In general, the
test is more stringent as density and viscosity decreases with various products.
Alcohol-based fuels and bio-blended fuels are appreciably dissimilar to petroleum-based fuels
without alcohol; the evaluation must specifically test using a representative product and under
reasonable conditions likely encountered in the field, such as presence of water from such
common sources as tank top sumps. Considerations such as water miscibility with the fuel
blend, especially with alcohol-blended fuels, will require testing the various sensors or functions
of the TTT, tank-related, and other UST system component release detection methods, as
applicable.
Because water is essentially immiscible in petroleum-based fuels, a very small addition of water
to an UST storing petroleum-based fuels will cause a water phase to settle in the bottom of the
tank. This makes it relatively simple to determine the presence of water in USTs storing
petroleum-based fuels. However, low alcohol-blended fuels can hold approximately 0.5 percent
of water before phase separation occurs. As fuel temperature is lowered, the amount of water
needed before phase separation occurs is also lowered. Because water alters the solubility of
alcohol in gasoline, when phase separation occurs in E10, for example, the separated phase
consists of an ethanol-water mixture with a density greater than ethanol but less than water. If
water entering an UST does not mix into a low alcohol-blended fuel, a separated aqueous phase
will collect at the bottom of the UST. However, once the UST receives a fuel drop and the
contents mix, the water is absorbed into the fuel until it reaches saturation.
As mentioned previously, water absorbed into alcohol-blended fuel will also increase the density
of the alcohol blend as well as other physical parameters, thus making proper selection of
volumetric correction factors difficult. In addition, a certain amount of water can be absorbed in
ethanol without an increase in volume and without separating at the bottom of the tank. In a
large volume of stored fuel, the amount of water absorbed into the alcohol fraction of an alcohol-
blended fuel could be appreciable and undetected.
Given the variability of the proportion of bio-components in fuels during testing, evaluators
should analyze the true proportion of alcohols such as ethanol or biodiesel to petroleum fuel and
record their findings with the test results. Following the ASTM International standard methods
presented below or another national voluntary consensus code, analyze an aliquot of the fuel for
the biofuel content. This is to characterize the fuel for testing and listing the method. Table 1
below specifies the methods that may be used for bio-component analysis by fuel.
10

-------
Table 1. Analytical Methods For Bio-Component Determination
Method
Designation
Method Title
Fuel Product
ASTMD7371
Determination of Biodiesel (Fatty Acid Methyl Esters)
Content in Diesel Fuel Oil Using Near Infrared Spectroscopy
Biodiesel
ASTMD4815
Standard Test Method for Determination of MTBE, ETBE,
TAME, DIPE, tertiary-Amyl Alcohol and CI to C4 Alcohols
in Gasoline by Gas Chromatography
Alcohol blend
up to 20%
ASTMD5501
Standard Test Method for Determination of Ethanol and
Methanol Content in Fuels Containing Greater than 20%
Ethanol by Gas Chromatography
Alcohol blend
over 20%
3.4 Leak Simulation Equipment
The test procedures call for inducing leaks in the tank. The method of inducing leaks must be
capable of being used with the release detection method. The test design in Section 4 gives the
nominal leak rates that the evaluator should use. These leak rates refer to leak rates that would
occur under normal tank operating conditions. The required leak rates for an evaluation on the
primary space of a tank will also apply when performing an evaluation on a method designed to
perform tightness testing on areas of the tank other than the primary space such as a tank
interstice and pipeline interstice.
The evaluator must disclose and validate the method used in the evaluation for simulating leaks.
While it may not be possible to achieve the nominal leak rates exactly, the method used to induce
the leak rates should be capable of being reasonably close to the nominal rates. Maintain and
record the induced leak rates. Compare the leak rates measured by the release detection method
to the induced leak rates.
Although certain leak simulation approaches may work for some non-volumetric methods, most
methods will require a means of simulating leaks that is adapted to their specific principle of
operation. It is the evaluator's responsibility to determine the appropriate leak simulation
approach and adequately determine the performance of each specific tightness test method.
3.4.1 Leak Simulation Approach For Non-Volumetric Methods To Include Acoustical
And Pressure-Vacuum Decay Methods
Some commercially available methods are based on corresponding pressure changes and
acoustical signals generated when product flows through an orifice or when air is drawn through
an orifice or hole in the tank that allows it to leak. In order to simulate a leak condition for such
a method, you must introduce an orifice into the tank so that product or air can flow through it
during the test. The orifice must be calibrated using diesel fuel with the equivalent pressure
exerted by the weight of diesel fuel in an 8-foot column to the desired leak rate of 0.10 gal/hr.
For desired leak rates other than 0.10 gal/hr for containment sumps, the orifice must be
11

-------
correspondingly calibrated using diesel fuel with the equivalent pressure exerted by the weight of
diesel fuel in an 8-foot 3column to the desired leak rate.
Orifice Calibration
You may use a variety of components to create an orifice such as a syringe, precision drilled
holes, a capillary tube or other devices that can be calibrated to allow liquid to flow through at
the desired rate. Calibrate the orifice to a leak rate of 0.10 gal/hr or less using diesel fuel with
the equivalent pressure exerted by the weight of diesel fuel in an 8-foot column to the desired
leak rate of 0.10 gal/hr. For desired leak rates other than 0.10 gal/hr for containment sumps, the
orifice must be correspondingly calibrated using diesel fuel with the equivalent pressure exerted
by the weight of diesel fuel in an 8-foot column to the desired leak rate. For non-volumetric test
methods, list the test method as capable of detecting leaks in liquids that are more viscous than
the liquid used to calibrate the orifice, if the evaluator determines that no other factors need to be
considered relative to the method's ability to detect a leak. The vendor may also choose to
perform additional testing on tanks containing more viscous liquids using orifices calibrated
accordingly. The same orifice calibration procedure applies to evaluating modes or methods of
leak detection equipment intended for use in the wetted and non-wetted portions of the test tank.
3.4.2 Leak Simulation Approach For Tracer Methods
Two types of leak simulation equipment are required, depending on the type of tracer technique
you are testing.
•	For methods that rely on detecting the loss from the tank of product containing tracer, the
simulation equipment must be capable of delivering a liquid containing the tracer into the
backfill close to the tank. The rate of delivery controls the volume of product introduced
in the backfill.
•	For methods that rely on detecting the loss of gaseous tracer from the tank, the simulation
equipment must be capable of delivering the tracer gas into the backfill in known
quantities so the evaluator can evaluate the ability of the method to detect the tracer in the
backfill.
In either case, the amount of tracer introduced into the backfill should reflect the amount that is
released if the tank leaks at a rate of 0.10 gal/hr or less. To do this, the rate of delivery controls
the amount of material introduced into the backfill. To simulate a 0 leak rate, introduce the
tracer material into the test tank and mix with the product as appropriate. However, you can
instead introduce a blank spike without a tracer into the backfill.
When testing tracer methods, additional considerations apply. While petroleum products spiked
with tracer are ideal, introducing regulated products into the ground is prohibited in almost all
situations. Therefore, for test purposes, the carrier for liquid tracers should be a non-regulated
liquid such as vegetable oil. Evaluate the concentration of tracer in the carrier to reduce the
actual volume of material introduced into the ground. The evaluator must separately determine
that the tracer is readily soluble in the regulated product.
12

-------
The evaluator can use direct injection of the tracer gas diluted in air to evaluate methods, which
rely on the loss of tracer gases from the tank. The concentrations of tracers injected during the
simulation process should approximate those contained in the tank during an actual test.
Other non-volumetric methods may use physical or chemical principles different from those of
the methods in these examples. The evaluator must develop a method of leak simulation that is
appropriate for the specific test method.
3.4.3 Leak Simulation Approach For Tightness Test Methods Using An Optical Device
For a tightness test method that uses an optical device such as a camera or a laser, the evaluator
must develop a method of leak simulation that is appropriate for the specific test method. For
optical devices, it is most likely that the method is a non-volumetric tank tightness test method
which requires a total of 42 tests. Of these 42 tests, perform 21 tests with the tank in a tight
condition and perform 21 tests with the tank simulating a 0.10 gal/hr or less leak. Tightness test
methods with an optical device may include methods that look for a leak inside the primary
space of a double-walled tank filled with dyed brine and contrasts with the color of the tank wall
or in a containment sump. The evaluator must determine an appropriate leak simulation
approach to ensure the method is properly evaluated.
Some optical tightness test methods, such as a camera based system, have a specified wait period
where the camera system is simply waiting for any visible leak to develop within a specified
amount of time, for example 60 or 120 minutes. Once the appropriate leak volume is simulated
and blind to the vendor, the vendor can then perform the test to determine whether the tank or
containment sump is tight or leaking. If an orifice is used, calibrate the orifice using diesel fuel
with the equivalent pressure exerted by the weight of diesel fuel in an 8-foot column to the
desired leak rate of 0.10 gal/hr. For desired leak rates other than 0.10 gal/hr for containment
sumps, the orifice must be correspondingly calibrated using diesel fuel with the equivalent
pressure exerted by the weight of diesel fuel in an 8-foot column to the desired leak rate.
For a tank tightness test method with a camera in a new tank not yet in service, the evaluator
must take into account that many newly installed tanks may have ballast water at the bottom of
the tank. If the method is capable of detecting leaks in a tank with ballast water, then the
evaluation must include performing tests with and without ballast water present. If there is a
limit on the amount of ballast water present in the tank for the method to perform properly, then
the evaluator must note this limitation on the performance of the method. Leaks must be
simulated in several areas including the top, bottom, end, and sides of the inside of the tank in
order to ensure performance of the tightness test method is adequately evaluated on all areas of
the tank. The number of tests for a VTTT or NVTTT on optical devices is the same as other
tightness test methods, for example 24 for VTTT and 42 for NVTTT. If there are relevant
limitations on the tightness test method found during the evaluation, the evaluator must note
those limitations.
13

-------
3.5 Sensor Evaluation Equipment
3.5.1	Liquid Sensor Test Vessel
The equipment to test a liquid sensor consists of a test vessel large enough to accommodate the
size of the sensor, typically a vertical cylinder standpipe. The height of the test vessel must be 8
inches or more. Product miscible with water must be treated or properly disposed at the
conclusion of the test. Minimizing waste is a consideration in determining the size of the test
vessel. Accurately measure the test vessel liquid height level to ± 0.001 inch. Mount the liquid
sensor in the same relation within the test vessel as it would be in the UST system.
In addition, the evaluator needs a means of repeatedly adding a small measured amount of liquid
such as product and water to the test vessel. You may establish the simulated ingress using a
variety of equipment; however, a peristaltic pump has been successfully used in the past. With
this approach, use an explosion-proof motor to drive a peristaltic pump head. Choose the
appropriate size of the pump head and tubing to provide the desired flow rates or liquid height
increase rate depending on the geometry of the test vessel and sensor. You should use a variable
speed pump head so you can achieve different flow rates with the same equipment. Direct the
flow through a rotameter so you can monitor and control the flow.
3.5.2	Vapor And Pressure Decay Test Vessel And Leak Simulation
The vapor test vessel equipment consists of compressed gas cylinders of test gases certified
accurate to ± 2 percent and ultrahigh-purity air, pressure regulators of 0 psi to 15 psi, tubing,
valves, tubing connectors, rotameter, test vessel, thermocouple of 0°C to 40°C to within ± 1°C,
and manometer at least 0-10 inches of water ± 5 percent. All of the equipment must be
constructed from materials that are inert with respect to the test gases. Use this equipment to
minimize the potential interferences of temperature changes, high temperature, excessive test
apparatus volumes, and leaks in the test vessel. Monitor the temperature and tests conducted at
normal laboratory temperatures. Maintain the internal pressure of the test vessel at a constant
pressure of ± 0.2 inches of water relative to ambient pressure. Keep the vessel volume as small
as possible without interfering with the operation of the sensor. There must be an inlet and an
outlet for flow of test conditions and fittings must allow connection to the sensor, a manometer,
and a thermocouple. See Figure 1 for an example schematic of a vapor test chamber.
14

-------
Figure 1. Example Schematic Of A Vapor Test Chamber
mm
Legend:
R - Regulator
C - Test Chamber
M - Manometer
P - Detector Probe
T - Thermocouple Probe
* - Used only with aspirating
detectors
3.6 Miscellaneous Equipment
As noted, the test procedures may require the partial emptying and filling of the test tank. One or
more large capacity fuel pumps may be required to accomplish the filling in a reasonably short
time. The evaluator may need hoses or pipes for fuel transfer. Some test methods require
reserve product for calibration or establishing a specified product level. In addition, containers
may be necessary to hold this product as well as that collected from the induced leaks. You may
need a variety of tools to make the necessary connections of equipment.
This procedure requires that before fuel is transferred to the test tank, a method of heating and
cooling the fuel must be provided, such as pumping the fuel through a heat exchanger or by
placing heating and cooling coils in the supply tank or tank truck.
15

-------
Section 4: Test Procedures
The evaluator can measure the overall performance of the method by comparing the method's
results of either leaking or tight tank to whether a leak was actually induced. The evaluator can
measure performance under a variety of realistic conditions, including temperature changes and
filling effects, if applicable. The evaluator is responsible for adding any other variables that may
affect a specific non-volumetric method. The range of conditions need not represent extreme
cases that might be encountered, because extreme conditions can cause any method to give
misleading results. If the method performs well under various test conditions, then it should
perform well in the field. Document the testing using the appropriate forms in Appendices B, C,
or D for volumetric, non-volumetric, and sensors, respectively.
The test procedures have been designed so the evaluator can perform additional statistical
analyses to determine whether the method's performance is affected by the size of the leak or
other factors. Conduct these additional analyses only if the method makes a substantial number
of mistakes so the proportion of errors is between 0 and 1 for some subsets of the data. Thus, the
additional analyses are only relevant if the method does not meet the performance standard.
The basic test procedures introduce two main factors that may influence the test: size of leak and
temperature effects.
•	Size of leak - Evaluate the method on its ability to detect leaks of specified sizes. If a
method cannot detect a leak rate of 0.10 gal/hr, which may vary for containment sump
testing, or if the method identifies too many leaks when no leak is induced, then its
performance is not adequate.
•	Temperature effects - Three temperature conditions should be used: added product at the
same temperature as the in-tank product, added product that is warmer than already in the
tank, and added product that is cooler. The temperature difference should be 10°F and
should be measured and reported to the nearest degree F. This establishes method
performance over a large temperature range and encompasses the range of seasonal
temperatures in many parts of the U.S. For some methods, the temperature difference is
needed to ensure the method can adequately test under realistic conditions. Compare the
performance under the three temperature conditions to determine whether these
temperature conditions affect the method's performance. Note that some non-volumetric
methods require an empty tank or do not require a specific product level. If the principle
of the non-volumetric method is not affected by product temperature as determined by
the evaluator, the test need not include this set of conditions, although the total number of
tests remains the same.
Non-volumetric test methods operate on a wide variety of physical and chemical principles.
Consequently, each method may have a different set of sources of interference related to its
operating principle. The evaluator must consider possible sources of interference for the test
method. Possible sources of interference might include noise, high water content, and turbidity.
The evaluator must report the list of the sources considered and his or her conclusions.
16

-------
The groundwater level is a potentially important variable in tank testing. The evaluator must, as
applicable, document the method's means of dealing with groundwater in the test results.
If the method uses a water ingress method to account for high groundwater levels, use EPA's
Standard Test Procedures For Evaluating Release Detection Methods: Automatic Tank Gauging
Systems. to evaluate two aspects of the method's water sensing function: the minimum
detectable water level and the minimum detectable change in water level. Together, you can use
these with the dimensions of the tank to determine the ability of the method's water sensor to
detect inflows of water at various rates.
Tank deformation may also be a potential interference the evaluator considers, especially if test
results are inconsistent across multiple test sites. Even with this potential, deformation may be
negligible and difficult to determine given other uncontrollable factors across different test tanks.
If one tank is used for the entire evaluation, deformation is controlled.
4.1 Environmental Data Records
In general, the evaluation test procedures require that the evaluator record the conditions during
the evaluation. In addition to all the testing conditions, document the groundwater level, if it is
above the bottom of the tank, and any special conditions that might influence the test results,
including weather changes.
When testing tracer methods, the evaluator should also document the tank environment as
completely as possible. Prepare a detailed site diagram, which identifies the positions of the
tanks, piping, and other features present at the site. Verify that the type of backfill and soil at the
site is, at a minimum, porous enough to allow migration of vapors from the leak to the sensors.
Do not conduct the evaluation under backfill conditions outside the range suggested by the
vendor. The range of conditions must be listed in the report.
The operating manual should describe both normal and unacceptable test conditions for each
method and should provide a reference against which the evaluator can compare the existing test.
Do not conduct the evaluation under conditions outside the vendor's recommended operating
conditions.
Record the following tank and product information, if applicable:
•	Type of product in tank;
•	Bio-component of product, if any;
•	Type of tracers, for example liquid or gas;
•	Tank volume;
•	Tank dimensions and type;
•	Amount of water in tank before and after each test;
•	Temperature of product in tank before filling;
•	Temperature of product added each time the tank is filled;
17

-------
•	Temperature of product in tank immediately after filling; and
•	Temperature of product in tank at start of test.
4.2 TTT Evaluation Test Procedures
4.2.1 Volumetric TTT Methods
Following the trial run in the tight tank, conduct a minimum of 24 tests at the leak rates
presented in Table 2 and similar to the example test design presented in Table 3 for volumetric
TTT methods. If the vendor chooses the option to evaluate their method on tanks connected by
siphon piping, an additional 24 tests can be performed following the test design in Table 3. In
Table 3, LRi denotes the nominal leak rates and Ti denotes the temperature differential conditions
used in the testing. These 24 tests evaluate the method under a variety of conditions with the
tank level at 90 percent or higher. If a method is to be evaluated at 50 percent, assuming the
performance has been proven adequate at 90 percent and higher, an additional 12 tests can be
performed at 50 percent. If a vendor chooses a method be evaluated at a level below 50 percent,
assuming the performance has been proven adequate at 50 percent and higher, an additional 12
tests can be performed at the lowest level the method is capable of achieving.
These number of test requirements also apply to an evaluation performed on a method for
tightness testing on a tank interstice or containment sump.
Leak Rates
Induce the following four nominal leak rates during the procedure to evaluate the method at 0.10
gal/hr leak rate.
Table 2. Leak Rates To Evaluate A Method At 0.10 gal/hr Leak Rate
English Units	Metric Units
(gal/hr)	(milliliters per minute)
0.0	0.0
0.05	3.2
0.10	6.3
0.20	12.6
Temperature Differentials
In addition, use three nominal temperature differentials between the temperature of the product
to be added and the temperature of the product in the tank during each fill cycle. These three
temperature differentials are -10, 0, and +10°F or -5.6, 0, and +5.6 degrees Celsius (°C). A
national survey on typical tank testing conditions1 found that for 57 percent of the tests, the
difference in temperature between the product in the tank and the newly delivered product was
less than 5°F and 86 percent of the tests for a temperature difference within 10°F of the product
1 Typical Tank Testing Conditions, Flora, Jairus D., Jr., and Jean Pelkey, Report on Work Assignment 22, Task 13,
EPA Contract No. 68-01-7383, Midwest Research Institute, December 2, 1988.
18

-------
in the tank. Using a temperature differential of 10°F simulates the majority of the typical tank
conditions.
Conduct the tests in sets, where the two tests in a set have the same temperature condition and
fill, but differ in the time since filling and may have different leak rates.
It may be unnecessary to perform thermal conditioning when performing an evaluation on a tank
tightness test method that tests dry tank related components such as a containment sump.
Randomization
Conduct 24 tests by inducing the 12 combinations of the four leak rates (LRi, LR2, LR3, and
LR4) and the three temperature differentials (Ti, T2, and T3) at the method-specified product
volume outlined in Table 3.
The evaluator is responsible for randomly assigning nominal leak rates of 0.0, 0.05, 0.10, and
0.20 gal/hr to LRi, LR2, LR3, and LR4 and nominal temperature differentials of -10°, 0°, and
+10°F to Ti, T2, and T3.
After a trial run, the evaluator or vendor will operate the method as it would be in a commercial
facility and record the data. The randomization is used to balance any unusual conditions and to
ensure the vendor does not have prior knowledge of the sequence of leak rates and conditions to
be used.
In summary, each test set consists of two tests. Conduct each set of tests using two induced leak
rates and one induced temperature differential, which is the temperature of product to be added
minus the temperature of product in tank. Each set of tests indicates the sequence in which the
product volumes in gal/hr will be removed from the tank at a given product temperature
differential.
Notational Conventions
The nominal leak rates of 0.0, 0.05, 0.10, and 0.20 gal/hr are randomly assigned LRi, LR2, LR3,
and LR4. While you may not achieve these exact figures in the field, these rates are targets that
should be achieved within ±30 percent.
During each test, measure the induced leak rates, denoted by Si, S2, ... S24, for each of the 24
tests. Compare these leak rates against leak rates obtained by the vendors.
Denote the leak rates measured by the TTT method during each of the tests by Li, L2, ... L24 and
correspond to the induced leak rates Si, S2, ... S24.
The subscripts 1, ... 24 correspond to the order in which the tests were performed; see Table 3.
For example, S5 and Ls correspond to the test results from the fifth test in the test sequence.
19

-------
Table 3. Leak Rate And Temperature Differential Volumetric Test Design (Example)

Test
Set
Nominal Leak
Rate
(LR in gal/hr)
Nominal Temperature

No.
No.
Differential (T in °F)
Trial run
-
-
0
0
Empty/Fill cycle





1
1
lr2
T2

2
1
LRi
t2
Empty/Fill cycle





3
2
lr3
t3

4
2
lr2
t3
Empty/Fill cycle





5
3
LRi
t3

6
3
LR4
t3
Empty/Fill cycle





7
4
lr3
Ti

8
4
LRi
Ti
Empty/Fill cycle





9
5
lr2
Ti

10
5
LR4
Ti
Empty/Fill cycle





11
6
LR,
T3

12
6
LRi
t3
Empty/Fill cycle





13
7
LRi
t2

14
7
LR4
t2
Empty/Fill cycle





15
8
LRi
Ti

16
8
lr2
Ti
Empty/Fill cycle





17
9
lr3
T2

18
9
lr2
t2
Empty/Fill cycle





19
10
LR4
t2

20
10
lr3
t2
Empty/Fill cycle





21
11
lr2
t3

22
11
lr3
t3
Empty/Fill cycle





23
12
LR,
Ti

24
12
lr3
Ti
20

-------
4.2.2 Non-Volumetric TTT Methods
Following a trial run in the tight tank, conduct a minimum of 42 tests for non-volumetric TTT
methods for single tanks as outlined in Table 4. If the vendor wants optional evaluation on tanks
connected by siphon piping systems, perform an additional 42 tests. Table 4 presumes that
temperature effects can interfere with the method. The tests on a tank system comprised of tanks
connected by siphon piping requires monitoring and reporting of temperature.
In Table 4, LRi denotes the nominal leak rates and Ti denotes the temperature differential
conditions. These 42 tests for single tanks evaluate the method under a variety of conditions.
The evaluator cannot establish the leak rate the same way for all TTT methods. Sections 3.4.1,
3.4.2, 3.4.3 and 3.5.2 describe a variety of leak simulation setups. As technology advances, other
devices may be developed to simulate releases.
Arrange the base 42 tests in 21 sets of two tests each. Table 4 shows a possible ordering of the
21 sets. The evaluator should randomly rearrange the order of the sets so the leak rates are blind
to the vendor.
Leak Rates
Of the 42 tests, conduct half under tight-tank conditions, that is, at a leak rate of 0.0 gal/hr.
Conduct the remaining 21 tests with under induced leak conditions with leak rates not exceeding
0.10 gal/hr. Typically, all of these induced leak rates would be the same. The test schedule in
Table 4 is an example of 21 tests at a 0.0 gal/hr leak rate (LRi) and 21 tests at non-0 leak rates of
LR2 (0.10 gal/hr). For testing at leak rates other than 0.10 gal/hr., the evaluator will select
comparable induced leak rates based on the target leak rate.
The most direct evaluation of a non-volumetric method uses only 0.0 and 0.10 gal/hr leak rates.
This, assuming the test results had at most one error at each leak rate, provides the needed
performance evaluation. However, a vendor may want to claim his method exceeds EPA's
performance standards and establish the probability of detecting a smaller leak, for example 0.01
gal/hr rather than 0.10 gal/hr, is at least 95 percent. In that case, two approaches are possible.
One is to use the smaller leak rate as the induced leak rate. However, if the nominal leak rate
selected is close to or less than the leak rate the method can actually detect with 95 percent
reliability, the testing may result in too many detection errors at that reduced leak rate. In order
to demonstrate the method meets the performance standards, run the 21 induced leak rate tests
again using a nominal leak rate larger than the example of 0.01 gal/hr, for example, 0.05 gal/hr.
With input from the vendor, the evaluator will select the most appropriate approach for the
evaluation.
Temperature Differentials, If Applicable
If temperature differential is important for the test method, use three nominal temperature
differentials between the temperature of the product to be added and the temperature of the
product in the tank during each fill cycle. These three temperature differentials are -10, 0, and
21

-------
+10°F (-5.6, 0, and +5.6°C). The temperature differential of 10°F is a minimum. You may use
larger differences. When temperature differences are used, calculate and report the actual
differences.
Randomization
Conduct 42 tests consisting of combinations of the two leak rates (LRi = 0.0 gal/hr, LR2 = 0.10
gal/hr) and the three temperature differentials (Ti, T2, and T3). Arrange the 42 tests in sets, with
each set consisting of two tests performed at the same temperature differential. However, the
leak rates within a set must be randomly assigned to the first or second position in the testing
order. Table 4 outlines the test schedule.
A randomization of the test schedule is required to ensure the testing is conducted blind to the
vendor. The evaluator is responsible for randomly assigning the leak rate of 0.10 gal/hr to LR2
and nominal temperature differentials of-10°, 0°, and +10 °F to Ti, T2, and T3, following the
sequence of 42 tests as shown in Table 4. In addition, the evaluator should randomly assign the
set numbers of 1 through 21 to the 21 pairs of tests. The vendor should not know which induced
leak rate is used or which temperature condition is present in advance. The vendor should test
for the induced leak rate based on the instrumentation and standard operating procedure without
knowledge of the induced conditions. Perform randomization separately for each method
evaluated. In Table 4, it is assumed all tests are done at a product level appropriate for the leak
detection method being tested. This level may be around 95 percent which is considered full, or
it may be a level required by the specific leak detection method, such as 60 percent of liquid.
For partially full tests, a supplemental test of the ullage area is recommended so the entire
portion of the tank normally containing liquid is tested.
In summary, each test set consists of two tests performed using two induced leak rates and one
induced temperature differential, which is the temperature of product to be added minus
temperature of product in the tank. Each set indicates the sequence in which the induced rates
are used to remove the product volumes in gal/hr from the tank at a given product temperature
differential. In some cases, for example, when a partial vacuum is applied to the tank, the
simulated leak will not actually remove product from the tank. In this case, the indicated rates
are those at which product escapes or is removed from the tank if the induced condition is
present under normal tank operating conditions.
Notational Conventions
The induced nominal leak rates are denoted by LRi = 0.0 gal/hr, LR2, = 0.10 gal/hr. While you
may not achieve these exact nominal leak rates in the field, these rates are targets that should be
established by a calibration process. The maximum must be no more than 10 percent greater
than the nominal 0.10 gal/hr.
Calibrate the leak rates induced for each of the 42 tests for each test series and denote the rates
by Si, S2, ... S42. Denote the results of each test by Li, ... L42, with each Li being either tight or
leaking. The Li may be coded numerically, for example, Li = 0 for tight and 1 for leaking, for
convenience.
22

-------
The subscripts 1, ... 42 correspond to the order in which the tests were performed; see Table 4.
For example, Ss and Ls correspond to the test results from the fifth test in the test sequence.
Table 4. Leak Rate And Temperature Differential Non-Volumetric Test Design (Example)

Test No.
Set No.
Nominal Leak
Rate
Nominal
Temperature
Differential
(T in 0 F)



(LR in gal/hr)
Trial run


0
0
Empty/Fill cycle
1
1
lr2
t3

2
1
LRi
t3
Empty/Fill cycle
3
2
LRi
t2

4
2
LRi
t2
Empty/Fill cycle
5
3
LRi
Ti

6
3
lr2
Ti
Empty/Fill cycle
7
4
lr2
t3

8
4
LRi
t3
Empty/Fill cycle
9
5
lr2
Ti

10
5
LRi
Ti
Empty/Fill cycle
11
6
lr2
t2

12
6
lr2
t2
Empty/Fill cycle
13
7
lr2
Ti

14
7
LRi
Ti
Empty/Fill cycle
15
8
lr2
t3

16
8
LRi
t3
Empty/Fill cycle
17
9
lr2
t3

18
9
LRi
t3
Empty/Fill cycle
19
10
LRi
t2

20
10
lr2
t2
Empty/Fill cycle
21
11
lr2
Ti

22
11
LRi
Ti
Empty/Fill cycle
23
12
LRi
t3

24
12
lr2
t3
Empty/Fill cycle
25
13
lr2
t2

26
13
lr2
t2
Empty/Fill cycle
27
14
lr2
t3

28
14
LRi
t3
23

-------
Table 4. Leak Rate And Temperature Differential Non-Volumetric Test Design
(Example) (Continued)

Test No.
Set No.
Nominal Leak
Rate
Nominal
Temperature
Differential
(T in °F)



(LR in gal/hr)
Empty/Fill cycle
29
15
LRi
Ti

30
15
lr2
Ti
Empty/Fill cycle
31
16
LRi
t2

32
16
LRi
t2
Empty/Fill cycle
33
17
LRi
t3

34
17
lr2
t3
Empty/Fill cycle
35
18
LRi
t2

36
18
lr2
T 2
Empty/Fill cycle
37
19
lr2
Ti

38
19
LRi
Ti
Empty/Fill cycle
39
20
LRi
t2

40
20
lr2
t2
Empty/Fill cycle
41
21
LRi
Ti

42
21
lr2
Ti
Tx is the monitored temperature of the system.
4.3 Implementation Of The Test Procedures
The first test is a trial run. Inform the vendor that you are conducting this test with a tight tank
stable condition. Report the results of the trial run along with any other data, but these results
not explicitly used in the calculations estimating the method's performance.
There are two purposes to this trial run. One is to allow the vendor to check the tank testing
method before starting the evaluation. As part of this check, the vendor should identify and
repair any faulty equipment. A second part is to ensure there are no problems with the tank or
the test equipment. Identify and correct common field problems such as loose risers, leaky
valves, and leaks in plumber's plugs.
The results also provide additional verification the tank is tight and provide a baseline for the
induced leak rates during the evaluation.
24

-------
Conduct the testing using a randomized arrangement of nominal leak rates and temperature
differentials as illustrated in Tables 3 and 4 above, unless the evaluator determines the filling or
temperature changes are irrelevant for the particular non-volumetric method. The time lapse
between the two tests in each set should not exceed 30 minutes and preferably be 15 minutes or
less. After each set of two tests, the test procedure starts anew with emptying the tank to less
than half-full, refilling, and stabilizing as necessary. The details of the testing procedure are as
follows.
Step 1	Randomize the test conditions - The evaluator randomly assigns nominal leak
rates LRi, LR2, LR3, and LR4. The evaluator also randomly assigns the
temperature differentials of-10°, 0°, and +10 °F to Ti, T2, and T3. Keep these
blind to the vendor performing the testing.
Step 2	Set up - If not already done, install the leak simulation equipment according to
the installation procedures in the tank, ensuring the leak simulation equipment
will not interfere with the test equipment.
Step 3	Trial run - Following the test method's standard operating procedure, fill the tank
to the test method's recommended level and at a minimum allow for the
stabilization period called for by the method. Product added should be the same
temperature as the in-tank product. Conduct a test on the tight tank to check the
system, such as tank and plumbing, and the method. Perform necessary repairs or
modifications identified by the trial run.
Step 4	Empty tank - Partially empty the tank to half-full. Fill with product at the
recommended temperature. The temperature differential will be Ti. Record the
date and time after completing the fill. Allow for the recommended stabilization
period, but not longer. Induce the appropriate leak condition.
Step 5	Conduct release detection method testing - Continue with the methods standard
operating procedure and conduct a test on the tank, using the method's
recommended test duration. Record the date and start time of the test. Perform
this test under the first nominal leak rate of the first set.
When the first test is complete, determine and record the calibrated induced leak rate, Si, and the
method's reported leak rate, Li. If possible, also record the data used to determine the leak
condition and the method of calculation. Save all data sheets, computer printouts, and
calculations. Record the beginning and ending dates and times of the test. Also, record the
length of the stabilization period. Appendices B and C provide individual test log forms for
reporting these data and environmental conditions.
Record the temperature of product in the test tank and temperature of the product added to fill
the test tank; if not recorded, document why not on the log. After adding product to fill the test
tank, record the average temperature in the test tank. One way to measure the temperature of
product in the tank is to use a probe with five temperature sensors spaced to cover the diameter
of the tank. Insert the probe or install it permanently in the tank; use the temperature readings of
25

-------
those sensors in the liquid to obtain an average temperature of the product. The temperature
sensors can be spaced to represent equal volumes or the temperatures can be weighted with the
volume each represents to obtain an average temperature for the tank.
Step 6	Change the nominal leak rate to the second in the first set - Repeat step 5. Note
there will be an additional period, which is the time taken by the first test and the
setup time for the second test, during which the tank may have stabilized. When
the second test of the first set is complete, again record all results such as times
and dates, induced leak rate and test result, temperatures, and calculations.
Step 7	Repeat step 4 - Change the temperature differential to the next in the test design.
Step 8	Change the nominal leak rate to the first in the second set - Repeat step 5. Record
all results.
Step 9	Change the nominal leak rate to the second in the second set, if it is different -
Repeat step 6. Record all results.
Step 10 Repeat step 4 - Change the temperature differential to the following one in the
test design.
Step 11	Repeat steps 5 through 9, using each of the two nominal leak rates of the third set.
Repeat steps 4 through 9, which correspond to two empty and fill cycles and two sets of two tests
until all tests are performed.
The operating manual should describe normal and unacceptable test conditions for each method
and provide a reference for comparing the existing test. Do not conduct the evaluation under
conditions outside the vendor's recommended operating conditions.
The evaluation must test all aspects of release detection the vendor claims that the method can
detect. Examples of these aspects include in high groundwater and in questionably porous soils.
At a minimum, the method must be tested to detect leaks from any portion of the tank that
normally contains product. If a water ingress method is used, refer to EPA's Standard Test
Procedures For Evaluating Release Detection Methods: Automatic Tank Gauging Systems. If
other sensors are used, test procedures are included in Section 4.5 of this document.
4.3.1 Application Of The Test Procedure To Acoustical Methods
One category of commercially available non-volumetric test methods is based on acoustical
principles. If the method relies on a person's hearing, within one year of testing, this method
requires that testers undergo a hearing test and their results are within the normal range, 0 to 20
decibels (http://www.hopkinsmedicine.org/hearing/index.html). Use Occupational Safety and
Health Administration regulation 29 CFR 1910.95 pertaining to occupational noise exposure for
guidance. Testers using the TTT method should have their hearing tested regularly. If the
26

-------
acoustical method uses an automated or computer-based detection method that does not rely on
human hearing, then hearing tests are unnecessary.
Acoustical methods use sensitive hydrophones to detect an acoustical signal from the tank.
Record this signal and analyze it to identify a specific characteristic associated with a leak. One
such method places the tank under a partial vacuum and investigates the acoustical signal for a
characteristic bubble signature, induced when air is drawn from outside the tank, through an
unobstructed hole in the tank wall, and into the liquid contained in the tank being tested. As
stated in Section 3.3, the evaluator and the vendor choose the liquid used during the evaluation,
but it must be capable of being used with the release detection method. Leaks in the ullage
portion of the tank are identified by a particular frequency or whistle of air entering directly into
the ullage space through holes in the tank wall. Another approach analyzes the acoustical signal
for a characteristic sound of fluid flowing out of an orifice in the tank.
While these methods are called acoustical, they typically have additional modes of detecting
leaks and are used in conditions of a high groundwater level. Generally, they rely on
identification of water ingress to detect leaks in the presence of a high groundwater level.
Acoustical methods can be used with a wide range of product levels in the tank. The temperature
of the product in the tank does not affect these methods. You do not have to consider the
sequence of temperature and filling conditions with these tests. Fill the tank to a level in the
range specified by the method. Generally, it is assumed that acoustical tests do the testing at a
single product level. If multiple levels are used, perform equal numbers of tests at each level.
To induce a leak for the acoustical methods, use a device that creates the same signal an actual
leak would create. One way is to use an orifice-type leak simulator per Patent No. 5,168,748.
This consists of a pipe inserted into the tank through one of the tank openings. The pipe is sealed
to the tank. The bottom of the pipe is fitted with a cap that contains a calibrated orifice to allow
product to leak into the pipe at the calibrated leak rate. The orifice must be calibrated using
diesel fuel with the equivalent pressure exerted by the weight of diesel fuel in an 8-foot column
to the desired leak rate of 0.10 gal/hr. For desired leak rates other than 0.10 gal/hr for
containment sumps, the orifice must be correspondingly calibrated using diesel fuel with the
equivalent pressure exerted by the weight of diesel fuel in an 8-foot column to the desired leak
rate. This simulator works for either type of acoustical signal. Flow of liquid through the orifice
produces the signal typical of liquid flow. If the tank is under partial vacuum, air is drawn into
the tank through the orifice below the liquid level and produces bubbles. You need a means of
closing the orifice so a 0 leak rate is induced and kept blind to the vendor.
Since temperature differential should not affect the acoustical methods, we simplified the
approach discussed earlier in this subsection. The steps refer to Table 4, with the understanding
there are no differences among Ti, T2, T3, and the partial emptying and refilling is unnecessary.
It is assumed that acoustical tests use a single product level; see above.
Step 1	Establish leak rates to be tested - If only a single non-zero leak rate is used, select
a leak rate between 0.0 and 0.10 gal/hr. If the vendor wants to establish a smaller
detectable leak rate, you may use a value of less than 0.10 gal/hr.
27

-------
Step 2	Randomize the test conditions - If only two leak rates - 0.0 and one other - are
used, randomly assign one of them to LRi and the other to LR2 as in Table 4.
Randomly rearrange the order of the 21 pairs of tests listed in Table 4. This
allows for additional randomization and provides better control on keeping the
induced leak rates blind to the vendor.
Step 3	Set up - Have the vendor set up the test method in the tank. Assemble and install
the required test equipment.
Step 4	Trial run - Following the test method's standard operating procedure, fill the tank
to the recommended level. Have the vendor conduct a test with a known 0 leak
rate and verify the method has been installed and is functioning correctly. This
also confirms the tank is still tight and is capable of being used with the test
method.
Step 5	Conduct release detection method testing - Induce the leak rate called for in the
randomization developed above. Have the vendor test the tank with this induced
leak rate and report the results. Record the calibrated induced leak rate and the
vendor's results as either tight or leaking. Record the environmental conditions
data and other ancillary data on the test logs; see Appendix B.
Step 6	When the first test is completed, change the leak rate to establish the second leak
rate called for in the randomized series, per Table 4. When this induced rate is
established, have the vendor test the tank. Record the environmental conditions
data. After the vendor completes the test, record the reported result and the
induced leak rate.
Step 7	Repeat step 6 until all 42 tests are complete.
In each case where the method declares a leak when the simulated leak is set to a tight condition,
the evaluator must check to ensure the leak simulator is in the off position. Conversely, in each
case where the method test result declares a tight condition and the simulated leak has been
established, the evaluator must confirm the calibration of the simulated leak equipment
immediately while minimizing actions that could impact the performance of the leak simulation
equipment. If the calibration of the simulated leak equipment is not as expected, discard the test.
As described in Section 5, the method can produce no more than one false alarm and still pass
the standard evaluation. Thus, if a second false alarm occurs in the test series, the method fails,
and you should terminate testing. Similarly, if only one non-zero leak rate is used, and if a
second mistake is made with that non-zero leak rate, the method fails. At the point where the
evaluator determines the method fails, consider concluding testing. However, you can continue
testing to provide added information to the vendor. If you used a leak rate of less than 0.10
gal/hr, starting the test series again with a leak rate closer to 0.10 gal/hr may result in the method
passing at that rate, but not at the smaller leak rate. If no errors occurred during 20 tight tank or
20 induced leak tests, the method passes.
28

-------
4.3.2 Application Of The Test Procedure To Tracer Methods
The technical requirements for using tracers are described in the release detection section of the
federal UST regulation on vapor monitoring; see 40 CFR 280.43(e). You must consider the
following requirements in evaluating tracer methods:
•	The backfill must be porous enough for a sensor to detect easily a low diffusion of
vapors.
•	The tracer must be volatile enough to produce vapor levels, which are detectable by the
monitoring device.
•	Groundwater, rain, or soil moisture must not interfere with operating the monitor.
•	Background contaminations must not interfere with detecting releases from the tank.
•	Optimize the number and positioning of the monitoring wells for detecting leaks from
any part of the system.
Although these requirements are for continuous vapor monitoring devices, when a tracer
technique is used as TTT, the requirements also apply. Accordingly, the test procedures consider
these factors when evaluating tracer techniques.
There are many variables present in external monitoring, which are difficult to predict or control.
These include the nature of the backfill material; moisture content of the soil; size of the
excavation; type of soil surrounding the excavation; groundwater level; position of a leak relative
to the sampling locations; and whether the method is aspirated or passive. In general, some
minimum threshold concentration of tracer must be reached before a signal is generated. The
lower the threshold, the more sensitive the method, but the more susceptible it will be to false
alarms.
For test methods that involve the loss of product from the tank, design the induced leak rates to
introduce the amount of tracer material into the soil that is released by leak rates of the specified
size over the test period. Methods that add liquid tracer to the product specify a concentration of
tracer in the product. Using a concentration of 10 ppm, a leak rate of 0.10 gal/hr, and a test and
waiting time after introducing the tracer into the tank of 24 hours, you can calculate the amount
of tracer that will be released. Release this amount during the leak simulation. One way to
accomplish this is to make samples of a carrier, such as vegetable oil with tracer in the
appropriate concentrations that can be introduced into the environment. Use these samples to
spike the ground at small rates, giving the same amount of tracer released by the specified leak
rates.
If the method uses gas tracers, they can be introduced into the ground to simulate leaks by using
a flow meter to allow the gas to flow at the rate that will occur under the in-field testing
conditions. For example, simulate a leak at 2 pounds per square inch (psi) and through an
appropriately sized orifice, representing a hole that leaks liquid product at the designated leak
rates of less than 0.10 gal/hr.
Note that once you introduce a tracer, gas, or liquid into the soil in a test, you must eliminate the
tracer before the next test. You may use forced air to disperse the tracer to levels that are not
29

-------
detected and interfere with the method; you may conduct the next test with a different tracer or
use a different site.
The following steps assume multiple tracers are available, one is used in the tank to investigate
the false alarm possibilities, and others are used in leak simulations. Neither the temperature
conditioning nor tank stabilization is an issue with tracer methods. Therefore, it is unnecessary
to change fuel temperatures and fill and empty the tank frequently as part of the evaluation. At
least 21 tests of the tank in the no-leak condition are required, as are at least 21 tests using the
induced leaks.
Step 1	Establish leak rates to be tested - Decide whether to use a single non-zero leak or
three non-zero leak rates and select these leak rates.
Step 2	Randomize the test conditions - Randomly assign the no-leak and leak conditions.
Randomly rearrange the order of the 21 pairs of tests in Table 4 that result from
assigning the leak rates.
Step 3	Prepare samples with carrier and tracer - Determine the rate of introducing tracer,
if a gas, or liquid carrier and tracer, if a liquid, into the backfill to simulate the
selected leak rates. If using a liquid tracer, prepare samples with the carrier and
tracer in the needed concentrations, label these with the randomized test sequence,
and provide them to the vendor. The vendor should not know whether or in what
concentration the tracer is in the leak simulation samples.
Step 4	Prepare the tank - If using a liquid tracer, have the vendor introduce it at the
desired concentration into the test tank and fill the tank to the desired level
following normal operating procedures for the method. If using a gas tracer,
empty the tank and have the vendor introduce the gas to the tank. The tank then
serves to provide data on the 0 leak rates.
Step 5	Locate sampling ports - Have the vendor locate the sampling ports. Also, have
the vendor locate a spiking port for leak simulation as far from the sampling ports
and as close to the tank as possible. Be careful not to damage the tank when
installing the ports in the backfill.
Step 6	Conduct the trial run - The trial run for a tracer verifies the method can be used
under certain site conditions. Introduce a compound at the spiking port. Sample
test locations to determine whether the compound is detected. The trial run
verifies the soil or backfill conditions allow the tracer to migrate from the tank to
the sensors; it determines the time needed for the migration and, thus, establishes
a test time.
Step 7	Conduct release detection method testing - Have the vendor conduct a test of the
tank for a 0 leak rate.
30

-------
Step 8	Begin testing using the first non-zero leak rate - Have the vendor conduct a test.
Note: If using two different tracers, it may be possible for the vendor to conduct
the test on the tank at a 0 leak rate and the induced leak test at the same time.
Step 9	After completing the tests in steps 7 and 8, record the induced leak rate, the
vendor's determination of tight or leaking, and the environmental conditions data
on the test log; see Appendix C.
Step 10 Ensure you can use the test site for a second leak test by removing the current
tracer or using a different one - Start the next induced leak rate as in steps 7 and 8
and have the vendor conduct another test. Record all results.
Step 11	Repeat step 10 until the test series is completed.
The vendor should be able to conduct tests on the tank containing the tracer repeatedly for the 0
leak rate tests. In conducting the repeated tests on the tight tank to estimate P(fa), repeat the
steps of adding tracer to the product, which is actually a carrier, and mixing the tracer in the
product, which is a carrier. The process of adding and mixing tracer is a likely cause of false
alarms because it could lead to inadvertent release of tracer into the environment and be
mistaken for a leak. The vendor should be able to simulate adding and mixing the tracer by
using tracer-containing product or carrier and handling it in the same manner as the tracer
solution.
Assuming that at least two tracers are available, you can run the tight tank tests and the simulated
leak tests simultaneously. Prior to start of the test, prepare and code the containers of carrier.
For each test, introduce the carrier sample in the spiking port. Half of them will contain tracer
and half will not. Each test will consist of introducing one tracer, for example type A, into the
tank and another sample, either a blank or tracer type B, into the spiking port. The vendor will
sample the soil gas and report on the presence of any detected tracer. A finding of tracer A is a
false alarm. A finding of tracer B when it was spiked is a correct detection. If using additional
distinct tracer compounds, this process will continue spiking tracer C. A finding of both tracer B
from a previous spike and tracer C from the current spike is a correct detection.
As described in Section 5, the method can record only one false alarm and still pass. If a second
false alarm occurs in the test series, the method fails, and the evaluator may recommend to the
vendor to end testing. Similarly, if using only one non-zero leak rate, and if a second mistake
occurs with that non-zero leak rate, the method fails. At the point where the evaluator
determines the method fails, you may conclude testing. If using a leak rate of less than 0.10
gal/hr, start the test series again with a leak rate closer to 0.10 gal/hr; this may result in the
method passing at that rate, but not at the smaller leak rate.
4.4 Testing Problems And Solutions
Inevitably, some tests will be inconclusive due to broken equipment, spilling of product used to
measure the induced leak rate, or other events that interrupt the testing procedure. Presumably,
31

-------
the field personnel can judge the validity of a test result. If a test is invalid, the following rules
apply.
Rule 1	The minimum number of tests must be either 24 or 42 for volumetric or non-
volumetric testing, respectively. If a test is invalid, report the reason and rerun
the test. Note that the number of tests assumes all tests are done at a single
product level. If using multiple levels, equal tests at each level are required and it
may be necessary to increase the total number of tests.
Rule 2	If method fails during the first test, meaning the first test of a set of two, and if the
time needed for fixing problems is less than 4 hours, then repeat that run.
Otherwise, repeat the empty and fill cycle, as well as the stabilization period.
Record all times.
Note: Report the average stabilization time or average time after introducing the
tracer on the Results Of U.S. EPA Standard Evaluation form. If the delay
increases this time noticeably, then rerun the test sequence.
Rule 3	If method fails during the second test, meaning after the first test in a set has been
completed successfully, and if the time needed for fixing problems is less than 4
hours, then repeat the second run. Otherwise, repeat the whole sequence of empty
and fill cycle, stabilization, and test at the given conditions.
Rules 2 and 3 apply only if the testing schedule requires temperature conditioning effects.
Otherwise, the time between tests is unimportant.
4.5 Sensor Evaluation Test Procedures
When testing sensors, the operating principle of the sensor drives test design, and fuel products
dictate the limitations of its use in UST systems. Since a sensor may perform differently when
in contact with various fuel products, such as ethanol blends or ethanol blends contaminated
with water, it is necessary to follow the appropriate procedures with a range of these blends and
calculate the specificity and accuracy with each blend; see below. Results will show an
operating range of hydrocarbon content or ethanol content that can be presented as limitations of
use in the reporting forms. Test the sensors in the types of liquids that they would be expected
to respond to under normal operating procedures. However, liquid level sensors should respond
to any liquid after the liquid level exceeds the threshold. If the evaluator finds that the sensor
does not respond to a particular liquid type, note this on the report forms. The thresholds may
vary slightly as the product density varies for float sensors. The evaluator determines which
blends and how many different blends to test. Testing procedures for sensor functionality in
systems with alcohol blends must include testing with a variety of amounts of water to
determine whether water interferes with performance of sensors designed to react to a change in
electric potential, such as capacitance and conductivity sensors. At minimum, the evaluator
must test the vendor's desired alcohol blend and that alcohol blend with three water mixtures:
80 percent alcohol blend and 20 percent water; 60 percent alcohol blend and 40 percent water;
and 30 percent alcohol blend and 70 percent water. For sensors that discriminate between
32

-------
hydrocarbons and water intended to be used in alcohol blends, to determine reliability and
accuracy of a sensor for a specific alcohol blend, the evaluator must evaluate the discriminatory
sensor both with alcohol blend fuel that is fully in solution with water, as well as with distinct
phase separation layer with neat gasoline on top. The sensor may only detect a certain layer or
layers.
The following performance parameters, which are defined below, are determined by these test
procedures. Report the data collected on these parameters, as applicable, on the forms and
tables in Appendix D.
•	Threshold, lower detection limit - The smallest product thickness that the detector can
reliably detect.
•	Precision, standard deviation - Agreement between multiple measurements of the same
product level.
•	Detection time - Amount of time the detector must be exposed to product or test
condition before it responds.
•	Recovery time - Amount of time before the detector stops responding after being
removed from the product.
•	Specificity - Types of products a sensor will detect.
4.5.1 Liquid Phase Sensor Test Procedures
Before performing an evaluation testing with a sensor, ensure the sensor is functioning and
properly calibrated. Properly calibrate all equipment making independent measurements during
testing and ensure the equipment is in working order.
You can evaluate liquid sensors within a clear glass test vessel with a sufficiently large inner
diameter to accommodate the sensor without being excessively wide. You will need a method to
measure the liquid height. A simple way is to use a ruler, graduated in millimeters, affixed to the
outside of the test vessel. Use an explosion-proof pump for the product ingress and a peristaltic
pump to deliver water into the test vessel. When using a fuel pump, you must use tubing that is
compatible with fuel. Secure the tubing in place so the liquids will flow along the side of the
container to the bottom without touching the sensor. The fuel and water ingress rates are set to
achieve a height increase rate of approximately 5 millimeters per minute (mm/min). Calculate
the rate of height increase by taking into account the volume displacement of the sensor in the
test vessel. Once the sensor and ingress lines are situated in the test vessel, cover the top of the
vessel to minimize volatilization.
Before initiation of testing, configure the test vessel as described above and insert the sensor
through the top of the test chamber. The sensor configuration with respect to the test vessel - for
example, suspended, vertically or horizontally resting on the bottom of the test chamber - will be
in concert with requirements of the vendor supplied literature and as close to intended field-
operating configuration as possible. Operate all sensors according to vendor-supplied operations
manuals and guidance including wiring, data collection, and maintenance. Additional measures
may be appropriate to simulate the operating environment of the sensor, for example wrapping
33

-------
the test vessel to minimize light with optical sensors. Record this adjustment to the test set on
the data collection form.
Collect the following data for sensor testing: test start time; sensor actuation time; liquid level
height at activation; test end time; and test recovery time for each test condition. Use these data
to calculate the metrics for the performance parameters of the test sensor.
Liquid Detection Test In Dry Space
The tests presented in this section simulate ingress of product or groundwater into a dry
environment or secondary containment, such as an interstice. Non-discriminating sensors will
respond to the presence of any liquid; however, incorporate at least three initial detection liquids
into the test design for the specificity calculation. Use an evaluator-chosen diesel fuel,
groundwater and gasoline or alcohol blend (as applicable), then perform 10 replicate tests on
each liquid. Use the most common gasoline blend or alcohol blend at the time of testing. For
testing sensors intended to be used in alcohol blends that are designed to react to a change in
electric potential such as capacitance and conductivity sensors, at minimum, the evaluator must
test the alcohol blend, in addition, that alcohol blend with three water mixtures: 80 percent
alcohol blend and 20 percent water; 60 percent alcohol blend and 40 percent water; and 30
percent alcohol blend and 70 percent water.
After placing the sensor inside the empty test vessel and activating it for data collection as per
the vendor instructions, monitor the output for 30 minutes as a blank test to establish the baseline
signal. Pump the product or groundwater from the graduated cylinder into the test vessel at
approximately 5 mm/min. Ensure the sensor is in place to detect this ingress of liquid and react
with a positive test result or not react in a negative test result.
At the completion of the tests, remove the sensor and the liquid from the test chamber. Measure
the liquid volume without the sensor and then handle the liquid appropriately by treating or
disposing of it properly. Rinse the sensor with water or clean it by following the vendor
recommended recovery procedure; monitor the recovery time.
With most liquid level sensors, conduct the following procedure with water, non-alcohol blended
gasoline, and diesel fuel. If the sensor is to be used with alcohol blends, test with the desired
alcohol blend. In addition, use the identified water, alcohol, and petroleum fuel mixtures
identified above for sensors designed to react to a change in electric potential. To determine the
threshold and precision of a sensor, follow the steps below.
Step 1	Set up - Mount the sensor in the test vessel with a known, uniform diameter from
top to bottom. Fasten the sensor securely so it is in contact via its normal
orientation with the liquid test vessel bottom.
Step 2	Blank test - Activate and monitor sensor for a minimum of 30 minutes to
establish a baseline.
34

-------
Step 3	Conduct liquid level testing - Add liquid such as product or water to the test
vessel from a burette capable of reading volume to the nearest 0.2 ml or pump
liquid into the vessel at a height increase of approximately 5 mm/min. Add liquid
until the sensor responds to the liquid. Allow adequate time between increments
for the sensor to respond if the response time is not instantaneous. Once the
sensor activates, the initial detection test is complete.
If the sensor does not activate, bring the liquid height to 20 percent higher than
the vendor stated actuation height and turn off the pump. Wait for 60 minutes
without detection before aborting the test. If a specific wait time is required, the
initial detection test is complete after the wait time has elapsed.
Step 4	When the approximate threshold has been determined, remove the sensor, and
empty the cylinder of liquid, then perform a repeat measurement.
Step 5	For subsequent measurements, add liquid quickly to just below the threshold
level.
Step 6	Add liquid very slowly until the sensor responds.
Step 7	Repeat steps 3 through 5 a minimum of 10 times for each liquid.
Step 8	Record all information in an appropriate manner.
The evaluator will determine if the sensor is affected by different product. To determine the
specificity of the sensor, use multiple products following the same test procedures to collect a
minimum of 10 replicates to compare the performance.
Test the sensors in the types of liquids they respond to under normal operating procedures.
However, liquid level sensors should respond to any liquid after the liquid level exceeds the
threshold and triggers the switch contact. If the evaluator finds that the sensor does not respond
to a particular liquid type, record this finding on the report forms. The thresholds may vary
slightly as the product density varies for float sensors.
Product Layer Detection On Top Of Water
Some external sensors are designed specifically not to alarm with water and to detect fuel
product. In this case, for non-alcohol blends, water may accumulate in the space and if fuel is
present, it will collect on top of the water creating a hydrocarbon layer. Install the sensor in the
test vessel as it is used with water that has a layer of product on it. Test the sensor 10 times at
each test product thickness of 0.0250 centimeters (cm) or 0.0625 in; 0.32 cm or 0.125 in; and
0.64 cm or 0.25 in on two different fuel products; note that these thicknesses are 1/16 inch, 1/8
inch, and 1/4 inch, respectively. Test diesel fuel and if the sensor is to be used with alcohol
blends, test with the vendor chosen alcohol blend. To determine reliability and accuracy of a
sensor for a specific alcohol blend, the evaluator must evaluate the discriminatory sensor both
with alcohol blend fuel that is fully in solution with water, as well as with distinct phase
35

-------
separation layer with neat gasoline on top. The sensor may only detect a certain layer or layers.
With input from the vendor, the evaluator chooses the product as to the intended purpose of the
sensor. With the collected data, the evaluator determines the accuracy and precision of the
sensor.
Visually, it is difficult to measure the thickness of the product layer on water, especially if the
product is somewhat miscible in water. For non-alcohol blends, the evaluator may determine the
layers mathematically under the assumption that there is no interaction of the product with water.
Where product is appreciably miscible in water for fuels such as alcohol blends this assumption
cannot be made. Regardless of type product, you must know the dimensions of the test vessel to
calculate the thickness of the product layer. Conduct random tests and allow no more than 24
hours for the sensor to react.
After completing each test, rinse the sensor with water or clean it by following the vendor-stated
recovery procedure; monitor the recovery time.
Step 1	Cross sectional area - Estimate the area of the cross section of the sensor that is
parallel to and at the same level as the test product. Calculate the cross-sectional
area of the test vessel.
Step 2	Set up - Mount the sensor in the test vessel. Securely fasten the sensor so it is in
contact via its normal orientation with the liquid test vessel bottom.
Step 3	Blank test - Calibrate the sensor. Activate and monitor sensor for a minimum of
30 minutes to establish a baseline. If the output is unstable after 30 minutes, wait
until it becomes stable.
Step 4	Add water - Add the appropriate volume of water, or approximately 2 liters, to
the test vessel as stated by the vendor. The volume of water added must allow the
sensor to be fully functional. The water should be within 2°C of room
temperature, which should be between 15°C and 28°C.
Step 5	Determine the amount of product to add to the water - Calculate the volume of
the product to add to the test vessel for each product layer thickness of 0.04 cm,
0.32 cm, and 0.64 cm with the following equations:
Volume (mL) = th x (ac — ad)
Where th is the desired product thickness in cm; ac is the test vessel cross
sectional area in cm2; and ad is the estimated sensor cross sectional area in cm2.
Begin testing with the thickest layer and continue to the smaller layers.
Step 6	Add the product and conduct the testing - Add the calculated volume for the
thickest layer to the test vessel without splashing or contacting the container
walls. Cover the test vessel immediately to reduce product loss. Do not stir or
otherwise disturb the test setup. Monitor the output of the sensor. For
36

-------
quantitative sensors, monitor the output at least until the signal becomes stable or
24 hours elapses, whichever is shorter. The period for detection time is from the
start time to the time the sensor reaches 95 percent of its final stable output. For
qualitative detectors, monitor the output until it activates or 24 hours elapse,
whichever is shorter. Record the results.
Step 7	Clean vessel - Remove the liquid from the test vessel and rinse the vessel with
water, then acetone to remove all product residue.
Step 8	Repeat steps 3 through 5 a minimum of 10 times for each product layer. Record
all results.
4.5.2 Product Vapor Phase Sensor Test Procedures
For gasoline, test the sensor with at least two gases. The first gas must be either benzene or 2-
methylbutane. The evaluator, with input from the vendor, will select the second gas. Include a
justification for the chosen gas in the test results. For fuel types other than gasoline, the
evaluator must select the test gases, with input from the vendor to closely match the fuel type for
which the sensor will be used. Include a justification for the chosen gasses in the test results.
The test gas concentrations by volume are nominally 0.005 percent of the lower explosive limit
(LEL), 0.025 percent of the LEL, 0.05 percent of the LEL, and 0.01 percent of the LEL. For fuel
types where test gas LEL is not applicable, the test gas concentrations by volume are nominally 5
percent of current test gas immediately dangerous to health or life (IDHL) value, 25 percent
IDHL, 50 percent IDHL, and 100 percent IDHL. Randomly conduct the tests 10 times at each
concentration for both gases. Allow the sensors up to 24 hours to respond to the test conditions.
When testing the specificity of the sensor, test multiple gasses with 10 replicates at one
concentration of 0.005 percent of the LEL or 50 percent IDHL where applicable. For gasoline
the suggested gasses are benzene, n-butane, n-hexane, isobutane, 2-methylpentane, 3-
methlpentane, and toluene. The evaluator, with input from the vendor, may test other gases.
After completing each test, purge the vessel with high purity air and measure the time until a
steady background or recovered response from the sensor is established. The nature of the
recovery response depends on whether the detector gives a quantitative or qualitative response.
Step 1	Randomize the test conditions - Randomly arrange the order of the gases at the
various concentrations for the test design.
Step 2	Set up - Install and calibrate the sensor into the vapor test vessel as stated by the
vendor in relation to how it is installed at an UST facility. Calibrate all
monitoring equipment. The seal between the sensor and the test vessel should be
gas tight.
Step 3	Purge the test vessel for at least three minutes with ultrahigh-purity air at 0.2
L/min before each test.
37

-------
Step 4	Introduce gas and conduct the testing - Add the test gas to the test vessel at the
sufficient rate to produce a 0.2 L/min test system vent flow rate according to the
random test design. Monitor the output of the sensor. For quantitative sensors,
monitor the output at least until the signal becomes stable at ± 2 percent of full
scale over 1 minute or 24 hours elapses, whichever is shorter. The period for
detection time is from the start time to the time the sensor reaches 95 percent of
its final stable output. For qualitative detectors, monitor the output until it
activates or 24 hours elapse, whichever is shorter. Record the results.
Step 5	Repeat steps 3 and 4 a minimum of 10 times for each gas and concentration.
Record all results.
4.5.3 Test Procedures For Tightness Testing Using A Vacuum Monitor On A Double-
Walled Tank Interstice With Or Without The Addition Of A Liquid Sensor
This evaluation determines the ability of a leak detection system to detect an air leak, fuel leak,
or water leak in the interstitial space of a double-walled tank. Results of this evaluation
determine whether the system can detect a leak of 0.10 gal/hr in addition to the time required for
the method to detect the induced leak. The leak must be calibrated using diesel fuel with the
equivalent pressure exerted by the weight of diesel fuel in an 8-foot column to the desired leak
rate of 0.10 gal/hr. For desired leak rates other than 0.10 gal/hr for containment sumps, the
orifice must be correspondingly calibrated using diesel fuel with the equivalent pressure exerted
by the weight of diesel fuel in an 8-foot column to the desired leak. The evaluation consists of
testing both the vacuum sensor as well as testing the liquid sensor. When testing the vacuum
sensor, use a test vessel of approximately 5 gallons to simulate the open space of a double-walled
tank interstice where air ingress leaks are induced. If there is a liquid level sensor present at the
low point of the interstitial space, there is no need for the vacuum sensor to be tested with
product leaks since the liquid sensor will alarm with the presence of product or water. When
testing the liquid sensor portion of the method, evaluate the liquid sensor following the liquid
phase sensor test procedure requirements listed in Section 4.5.1.
If the leak detection method does not include a liquid sensor in the interstice, it will be necessary
to perform additional tests with a simulated liquid leak into the test vessel to determine the
methods ability to detect a liquid leak. The leak, which can be generated using a variable valve
flow or an orifice, must be calibrated using diesel fuel with the equivalent pressure exerted by
the weight of diesel fuel in an 8-foot column to the desired leak rate of 0.10 gal/hr. For desired
leak rates other than 0.10 gal/hr for containment sumps, the orifice must be correspondingly
calibrated using diesel fuel with the equivalent pressure exerted by the weight of diesel fuel in an
8-foot column to the desired leak rate.
After the test vessel is set up with the leak detection method, perform a baseline test to ensure
there are no leaks in the system. Once the test vessel with the method is confirmed to be tight,
begin the evaluation.
If the method is not an automated system, the vendor will specify the parameters that indicate a
leak is detected.
38

-------
During the vacuum sensor portion of the evaluation, perform 21 tests with the vessel in the
non-leaking tight condition and 21 tests with an air ingress leak while inducing the pre-
calibrated leak. Regardless of the vacuum level the vendor uses, the induced leak will not
change from its pre-calibrated rate of 0.10 gal/hr using diesel fuel with the equivalent pressure
exerted by the weight of diesel fuel in an 8-foot column.
Vacuum Sensor Test Procedure
Step 1	Set up - The evaluator must set up the test vessel and verify it is in a tight non-
leaking condition. After verifying the vessel's tightness, then calibrate the leak at
a rate of 0.10 gal/hr using diesel fuel with the equivalent pressure exerted by the
weight of diesel fuel in an 8-foot column. Once the leak is calibrated, install the
leak simulation device onto the test vessel. The vendor then installs and calibrates
their sensor or vacuum gauge into the test vessel. Calibrate all monitoring
equipment. Perform the tightness test as specified by the vendor.
Step 2	Conduct the testing - If the leak detection method includes the use of a liquid
sensor, then perform a total of 42 tests. Of the 42 test total, perform 21 tests
with a leak induced at the pre-calibrated leak rate of 0.10 gal/hr and perform 21
tests in a tight condition. If the leak detection method does not contain a liquid
sensor, then 21 tests will need to be performed with each liquid the method
might encounter including water, unleaded fuel and diesel fuel. For the leak
induced tests, induce the calibrated leak allowing air or liquid to flow into the
test vessel until reaching the specified alarm level of vacuum. Close the air inlet
after the vacuum has decayed to the vendor specified level. Record the elapsed
time from when the leak was induced until the system alarms or the specified
alarm level of vacuum is reached. If the method fails to identify the tight or
non-tight condition within the vendor's specified time frame, the method fails
the evaluation test.
In each case where the method's test result declares a leak when the simulated
leak is set to a tight condition, the evaluator must confirm the test vessel is not
leaking. Conversely, in each case where the method's test result declares a tight
condition and the simulated leak has been established, the evaluator must
confirm the calibration of the simulated leak equipment. If the calibration of the
simulated leak equipment is not as expected or if a leak is found in the test
vessel, then discard the test. If a method incorrectly reports either a leak or tight
condition and all of the equipment is operating as it should, then the method
fails the evaluation test.
Step 3	Reestablish the vacuum - Reconnect the vacuum source and repeat the test by
establishing the normal operating level of the vacuum in the test vessel.
Step 4	Repeat steps 2 and 3 a minimum of 21 times with an air leak induced, 21 times
with no leak induced and, if required, 21 times with a liquid leak induced for any
39

-------
liquid that the method may encounter including water, unleaded fuel and diesel
fuel.
4.5.4	Recovery Time
After the end of the individual tests, turn off the pumps and remove the sensor from the chamber.
Rinse the sensor with water or follow the vendor-stated recovery procedure and monitor for
recovery time.
Step 1	After individual test, remove the sensor from the test condition and follow the
vendor-stated recovery procedure.
Step 2	Using a stopwatch or console, record the time required for the sensor to stop
alarming after the alarm condition is reported. If the sensor results are an output
signal, the recovery time is concluded when the sensor returns to within 5 percent
of the baseline level.
Step 3	Repeat the above procedures a total of 42 times with 21 air leaks induced and 21
tests in the tight condition. If a liquid sensor is not included in the method, repeat
the above procedures an additional 21 times for each type of liquid leak that the
method might encounter including water, unleaded fuel and diesel fuel. Perform
the leaks induced and the tight tests in a randomized order. Record the data on
the individual test logs.
4.5.5	Test Procedures For Tightness Testing On A Liquid Filled Interstice Of A Double-
Walled Pipeline Using A High Pressure And A Low Pressure Limit Switch Sensor
This section describes how to conduct testing on liquid filled interstitial monitors for double-
walled pipelines. You may apply the results of this evaluation to any leak detection system that
performs tightness testing on a pressurized liquid filled interstice of a double-walled pipeline
using high and low pressure limit switches.
Since interstitial monitoring systems are highly dependent on the type of piping materials used,
the performance of the tightness test method only applies to the type of pipeline and interstitial
fluid used during the evaluation.
The evaluation procedures during this evaluation include 42 tests with 21 tests performed with
the interstice in a non-leaking condition and 21 tests performed with a 0.10 gal/hr leak present.
Calibrate the leak to 0.10 gal/hr at the pressure level that the method's high pressure limit switch
uses.
One potential concern to evaluate for the method, other than the 42 leak performance tests, is
making sure the thermal effects do not cause the pressure level to fluctuate beyond the pressure
limit switches used to monitor for a leak. Two possibilities are: the pressure will drop too far
and trigger a leak alarm or the pressure will rise too high and trigger a high pressure alarm. In
40

-------
addition to the 42 tightness tests, consisting of 21 leak and 21 tight, also perform tests to
demonstrate the method can overcome thermal effects.
Applicability
Use this procedure to test any doubly contained pipeline system with a pressurized liquid filled
interstice using a high and low pressure limit switch. Connect the sensors to some type of
control panel that can be configured to provide the operator with an alarm or will shut down
dispensing, if a leak occurs. If the system also uses a liquid sensor in any way, evaluate the
liquid sensor following the liquid phase sensor test procedure requirements in Section 4.5.1.
Test Apparatus
Construction Of Test Line
To conduct these tests, construct or identify a pipeline with a known volume. You may perform
the evaluation in any volume line up to 50,000 gallons. The results of the evaluation are
applicable to all smaller pipelines of the same construction; therefore, the larger the test line, the
broader the applicability of the evaluation. The results are also applicable to larger pipelines of
the same construction with the restriction the lines be no more than 100 percent larger in capacity
than the test pipeline. The test pipeline must include one of each of the types of fittings normally
found in a service station and may be clustered together. Access to the ends of the test pipeline
must be provided for inducing leaks, circulation of fluid through the primary pipeline, or other
activities associated with the testing. The testing may be conducted in a laboratory or shop
environment.
You can use water as the liquid in the primary space of the pipeline for all the tests performed for
the evaluation. The liquid in the interstice must be of the same type used by the manufacturer for
installed systems. Fill the interstice using the same procedures as specified by the pipeline
manufacturer at a field installation or when pre-filled at the factory before shipping. This could
include gravity feed, evacuation of the interstice prior to filling, or other technique designed to
minimize the amount of air trapped in the interstice. When completed in a laboratory
environment, insulate the laboratory line from the environment so that temperature of the system
is not subject to rapid temperature fluctuations produced by the ambient conditions. You may
use aluminized mylar bubble pack or other easy-to-handle material.
Test Equipment
Heating And Cooling
Provide for circulating hot and cold water through the primary pipe during the evaluation
process. You can accomplish this by using the equipment described below or by another
equivalent method that can maintain the circulation of water at a constant temperature for one
hour or until the entire test assembly has reached thermal equilibrium.
Use an insulated 55-gallon drum or other suitable container as a reservoir. Lower the water
temperature to a nominal temperature of 32°F by adding crushed ice to the reservoir. If an
excess of ice is present, the temperature will be maintained at near 32°. Use a small, low-
41

-------
pressure pump to circulate the water through the primary pipe. The capacity of the pump must
be sufficient to provide a water flow rate between 5-10 gallons per minute.
Heat the water by using a small flow through heater in the water return line. The heater must be
capable of heating the water to at least 110°F and maintaining the temperature at 100°F during
the circulation.
Pressurizing The Pipeline
To provide for the pressure testing, use a pump capable of delivering a pressure up to the pipeline
system manufacturer's pressure limit, but not to exceed 45 psi. Connect the pump to the primary
line at either the inlet or outlet of the test assembly.
Induced Leaks
Calibrate the leak to 0.10 gal/hr at the pressure level that the method's high-pressure limit switch
uses. Induce leaks using any type device that adjusts with pressure fluctuations such as an
orifice or variable valve flow meter.
Temperature Measurements
Temperature measurements should be made to 0.5°F using a temperature device with an
accuracy of 0.5°F. The accuracy is less important than the resolution but calibrate all
temperature devices to within 0.5° of each other. Take temperature measurements in the
circulation reservoir and on the outside of the interstice under the insulation within 12 inches of
the inlet to the primary pipe.
Pressure Measurements
Make pressure measurements to 1.0 psig or better. The pressure gauge should have a range of
twice the expected pressure range of the testing and have an accuracy of at least 3 percent of full
scale.
Evaluation Procedures
Primary Pipe Pressure And Thermal Effect Test
Several types of tests must be conducted to establish the characteristic of the pipeline under
consideration. These include:
•	Effects of pressure in the primary pipe on the interstitial pressure
•	Effects of temperature on the pressure level in the interstice
•	Effects of a catastrophic failure of the primary pipe
•	Flow through the interstice
Conduct these tests by monitoring the pressure level in the interstice with a device capable of
measuring the actual changes produced in the testing. The high and low pressure limit switch
sensors used for monitoring cannot be used for these tests. The test procedures are summarized
below.
42

-------
Effects Of Pressure In The Primary Pipe On The Liquid Level
This test involves raising the pressure in the primary pipe from 0 psig to 45 psig. Monitor the
liquid level in the reservoir at regular intervals during this time.
Step 1	Set the pressure level in the interstice at the level specified by the vendor while
the pipeline temperature is at the ambient temperature of the testing facility.
Step 2	Raise the pressure in the inner pipe from 0 up to the maximum pressure specified
by the piping manufacturer, but not more than 45 psi. Raise the pressure in 15 psi
increments.
Step 3	Note the change in the interstitial pressure after each increment. This pressure is
approximately 1.5 times the pressure expected at a typical service station
installation.
Step 4	Hold the pressure at the highest pressure for at least 10 minutes before making the
final level measurement.
Step 5	Return the line to ambient pressure and note the final pressure in the interstice.
If the pressure increase of the primary pipe causes the pressure in the interstice to exceed the
high pressure limit switch, then the method fails the evaluation.
Effects Of Temperature On The Pressure Of The Interstice
This test involves circulating hot and cold water at a constant temperature through the primary
pipe. Measure the temperature of the interstice by placing a thermocouple between the bubble
pack insulation and the outer pipe. Measure the liquid level in the reservoir periodically during
the circulation until attaining a constant interstitial temperature and interstitial pressure. The
temperature of the circulated fluid should range from approximately 32°F, achieved by using ice
for cooling, to 100°F, which is a temperature range of approximately 68°F.
Conduct the test as follows:
Step 1	Circulate water at a nominal temperature of 32°F through the primary pipe for at
least 30 minutes. Maintain this temperature during the entire circulation period.
Step 2	Continue circulation until the interstice pressure is stable.
Step 3	Monitor the outer wall of the interstice with a thermocouple. If collecting data
manually, take data every 5 to 10 minutes until you obtain stable readings.
Step 4	Ensure the interstitial temperature measurement is stable before beginning the
temperature increase.
Step 5	Repeat this process using water heated to a nominal temperature of 100°F.
43

-------
Step 6	Continue circulation until the interstice pressure is stable.
Step 7	You can also conduct this process starting at the high temperature and going
down to the low temperature.
If the thermal effects cause the pressure in the interstice to exceed the method's high or low
pressure limit switch, then the method fails the evaluation.
Effects Of A Catastrophic Failure Of The Primary Pipe
Conduct the effects of a catastrophic failure of the inner pipe at a minimum of two locations.
The first should be within 3 feet of the liquid reservoir and the second at a point within 3 feet of
the far end of the test line. Produce the catastrophic leak by introducing the interstitial liquid
into the interstice at a pressure of 30 psi.
Step 1	Configure the test line to allow the introduction of interstitial liquid through a ball
valve and into the interstice at a pressure of 30 psi.
Step 2	Ensure the inlet for the catastrophic leak is within 36 inches of the reservoir for
one of the two tests.
Step 3	Rapidly open a valve capable of allowing a flow of at least 10 gal/min into the
interstice.
Step 4	Ensure the alarm system is capable of shutting off the turbine.
Step 5	Repeat this process at the far end of the pipeline.
If the high level limit switch is not triggered with the interstice pressured to 40 psi, or whatever
the evaluator deems appropriate based on the system's high level limit setting, then the method
fails the evaluation.
Tightness Test Evaluation Procedures On The Interstice
This evaluation procedure determines the ability of the method to detect a leak with a rate of 0.10
gal/hr or smaller in the interstice of a double-walled pipeline.
Step 1	Set up - The evaluator must set up the pipeline and verify that the interstice is in a
tight non-leaking condition. Fill the primary space of the pipeline with water.
Calibrate the leak simulation device to a leak rate of 0.10 gal/hr at the pressure
level to which the method's high pressure limit switch is set. Install the leak
simulation device at the furthest point away from the pressure limit switches.
Step 2	Conduct the testing - Perform a total of 42 tests. Of the 42 tests total, perform
21 tests with a leak induced at the pre-calibrated leak rate of 0.10 gal/hr and
perform 21 tests in a tight condition. For leak induced tests, induce the
calibrated leak allowing interstitial fluid to flow out of the interstice through the
leak simulation device until exceeding the specified alarm level of pressure.
44

-------
Shut off the leak simulation device once the pressure has decayed beyond the
vendor specified level. Record the elapsed time from when the leak is induced
until the system alarms or the specified alarm level of vacuum is reached. If the
method fails to identify the tight or non-tight condition within the vendor's
specified time frame, the method fails the evaluation test.
In each case where the method's test result declares a leak when the simulated
leak is set to a tight condition, the evaluator must confirm that the interstice is
not leaking. Conversely, in each case where the method's test result declares a
tight condition and the simulated leak is established, the evaluator must confirm
the calibration of the simulated leak equipment. If the calibration of the
simulated leak equipment is not as expected or if a leak is found in the interstice,
then discard the test. If a method incorrectly reports either a leak or tight
condition and all of the equipment is operating as it should, then the method
fails the evaluation test.
Step 3	Repeat the next test by establishing the normal operating level of the pressure in
the interstice.
Step 4	Repeat steps 2 and 3 a minimum of 21 times with a leak induced and 21 times
with no leak induced. Record all results.
45

-------
Section 5: Calculations
From the results obtained after testing volumetric and non-volumetric methods, evaluate the
method's performance through a series of calculations presented in Section 5.1 and 5.2,
respectively. If the method has more than one mode of release detection, then evaluate and
report the performance of the method for each testing mode separately. If the performance is
different for different modes, this may limit the conditions under which the method can be used
and report these under the limitations section of the results form.
After performing tests according to the schedule outlined in Section 4, a minimum of 24 or 42
test results will be available. If the P(fa) and P(d) are not at the regulatory level with the tanks
connected by siphon piping data included, the method is inappropriate for tanks connected by
siphon piping and is limited to single tanks. In this case, to test the tanks, break the siphon
connection in order to isolate each tank for separate testing.
5.1 Estimation Of The Volumetric Method Performance Parameters
After performing all tests according to the basic test design, a total of at least n = 24 data points
each of 4 leak rates x 3 temperature differentials completed twice of measured leak rates and
induced leak rates will be available. These data form the basis for the performance evaluation of
the method. Denote the measured leak rates by Li, ... Ln and the associated induced leak rates
by Si, ... Sn. Number these leak rates in chronological order. Table 5 summarizes the notation
used throughout this test procedure, using the example test plan of Table 3.
5.1.1 Basic Statistics
The number of tests is designated by n. Calculate the mean squared error (MSE), the bias (B),
and the variance of the method as follows.
Mean Squared Error, MSE
n
MSE = Y^li-Sd2/n
1 = 1
Where Li is the measured leak rate obtained from the ith test at the corresponding induced leak
rate, Si, with i =1, ... n.
Bias, B
n
B= ^ik-Sd/n
i=1
The B is the average difference between measured and induced leak rates over the number of
tests. It is a measure of the accuracy of the method and can be either positive or negative.
46

-------
Table 5. Notation Summary
Test No.
Pair
No.
Set
No.
Nominal
Temperature
Differential
(degree F)
Nominal
Leak Rate
(gal/hr)
Induced
Leak Rate
(gal/hr)
Measured
Leak Rate
(gal/hr)
Absolute
Leak Rate
Difference
|L-S|
(gal/hr)
1
1
1
t2
LRi
Si
Li
di
2
1
1
t2
lr2
s2
l2
d2
3
2
1
t2
lr4
s3
l3
d3
4
2
1
t2
lr3
s4
l4
d4
5
3
2
Ti
LRi
S5
L5
d5
6
3
2
Ti
lr4
s6
u
d6
7
4
2
Ti
lr2
s7
L7
d7
8
4
2
Ti
lr3
s8
L8
d8
9
5
3
t3
lr4
s9
L9
d9
10
5
3
t3
LRi
Sio
L10
dio
11
6
3
t3
lr3
Sn
Ln
dn
12
6
3
t3
lr2
Sl2
Li2
di2
13
7
4
t2
lr3
Sl3
Li3
di3
14
7
4
t2
lr4
Si4
Li4
di4
15
8
4
t2
lr2
S15
L15
dis
16
8
4
t2
LRi
Sl6
Ll6
di6
17
9
5
Ti
lr2
S17
L17
dn
18
9
5
Ti
lr3
S18
Ll8
dis
19
10
5
Ti
lr4
Sl9
L19
di9
20
10
5
Ti
LRi
S2o
L2o
d2o
21
11
6
t3
lr3
S21
L2i
d2i
22
11
6
t3
lr2
s22
l22
d22
23
12
7
t3
lr4
s23
l23
d23
24
12
7
t3
LRi
s24
l24
d24
Variance And Standard Deviation
Obtain the variance as follows:
n
Variance = ^[(Lj — St) — B]2/cLf
i=1
Standard deviation (SD) is the square root of the variance. Where df = degrees of freedom.
47

-------
Note: Plot the differences between the measured and induced leak rates against the time or the
order in which they are performed. This allows the evaluator to detect any patterns that might
exist, indicating potentially larger differences in the results from the first test of each set of tests,
among the three temperature differentials, or between in-tank product levels. This could suggest
the method calls for an inadequate stabilization period after filling, the method does not properly
compensate for temperature differences between in-tank product and product to be added, or the
method is influenced by the product level.
Note: Tank tightness tests usually require testing at 90-95 percent full. If a lower level is used,
the method is restricted to the lower level and you cannot use tanks that contain more product
than the level tested, unless the tank ullage was tested by another method.
It may also be useful to plot the differences between the measured and induced leak rates by
induced leak rate. This graphically shows the accuracy and precision of the method at the
various leak rates used during testing. See Section 5.3 for appropriate statistical tests.
Test For 0 Bias
To test whether the method is accurate - that is, the bias is 0 - perform the following test on the
bias calculated above.
Compute the t-statistic
tB = ^fnB/SD
From the t-table in Appendix A, obtain the critical value corresponding to a t with (n - 1) = df
and a two-sided 5 percent significance level. For 24 tests and 23 df, this t-value is 2.07.
Compare the absolute value of tB, abs(tB), to the t-value. If abs(tB) is less than the t-value,
conclude the bias is not statistically different from 0, and the bias is negligible. Otherwise,
conclude the bias is statistically significant from 0.
5.1.2 False Alarm Rate, P(fa)
Assume the normal probability model for the errors in the measured leak rates. Using this
model, together with the statistics estimated above, allows for the calculation of the predicted
P(fa) and P(d) of a leak of 0.10 gal/hr.
The vendor will supply the threshold (Th) for interpreting the results. Typically, the leak rate
measured by the method is compared to C and the results interpreted as indicating a leak if the
measured leak rate exceeds the vendor stated C. The P(fa) is the probability the measured leak
rate exceeds C when the tank is tight. Note that by convention, all leak rates representing
volume losses from the tank are treated as positive.
P(fa) is calculated by one of two methods, depending on whether B is statistically significantly
different from 0.
48

-------
P(fa) With Negligible Bias
In the case of a nonsignificant B in Section 5.1.1, compute the t-statistic
= Th/SD
where SD is the SD calculated above and Th is the method's threshold. Using the notational
convention for leak rates, Th is positive, obtain P(fa) from the t-table, using n-1 df. P(fa) is the
area under the curve to the right of the calculated value ti.
In general, t-tables are constructed to give a percentile, ta, corresponding to a given number of df,
df, and a preassigned area, alpha (a), under the curve, to the right of ta; see Figure 2. For
example, with 23 df and a = 0.05 (equivalent to a P(fa) of 5 percent), ta = 1.714.
In this case, however, the area under the curve to the right of the calculated percentile, ti, with a
given number of df needs to be determined. This can be done by interpolating between the two
areas corresponding to the two percentiles in Table A-l in Appendix A on either side of the
calculated statistic, ti.
The approach would be to use a statistical software package, for example, Microsoft Excel™,
SAS™ or SYSTAT™, to calculate the probability.
P(fa) With Significant Bias
The calculations are similar to those in the case of a non-significant B, except the B is included
in the calculation. Compute the t-statistic including B as follows:
f(t)
o
t
Figure 2. Student's t-Distribution Function
t2 = (Th — B)/SD
49

-------
P(fa) is then obtained by interpolating from the t-table, using n-1 = df. P(fa) is the area under the
curve to the right of the calculated value h. Note that Th is positive, but B can be either positive
or negative.
5.1.3 Probability Of Detecting A Leak Rate Of 0.10 gal/hr, P(d)
The P(d) with a leak rate of 0.10 gal/hr is the probability the measured leak rate exceeds Th
when the true mean leak rate is 0.10 gal/hr. As for P(fa), use one of two methods in computing
P(d), depending on whether the B is statistically significantly different from 0.
P(d) With Negligible Bias
In the case of a non-significant B - that is, the B is 0 - compute the t-statistic:
Next, using the t-table at n-1 = df, determine the area under the curve to the right of t3. The
resulting number is the P(d).
P(d) With Significant Bias
The calculations are similar to those in the case of a non-significant B, except the B is included
in the calculation. Compute the t-statistic.
Next, using the t-table at n-1 = df, determine the area under the curve to the right of t4. The
resulting number is the P(d).
5.2 Estimation Of The Non-Volumetric Method Performance Parameters
5.2.1 False Alarm Rate, P(fa)
Use the results obtained from the tests performed under tight tank conditions to calculate P(fa).
Let Ni denote the number of these tests. Let TLi denote the number of cases where the method
indicated a leak. If the test results, Lj, are coded as 0 when no leak is indicated and 1 when a
leak is indicated, then
where the sum is taken over the Ni tests at 0 leak rate. The P(fa) is estimated by the ratio
t3 = (Th- 0.10)/SD
t4 = (Th-B - 0.10)/SD
i=l
P(fa) = TLi/Ni
50

-------
In order for the method to meet the performance standards, the estimated P(fa) must be less than
or equal to 5 percent. Thus, in order for the method to meet the performance standards, TLi
must be no more than 1 if the standard number of tests are performed.
If the method did not identify the tank to be leaking when it was tight (TLi = 0), then the
proportion of false alarms becomes 0 percent. However, this does not mean the method is
perfect. The observed P(fa) of 0 percent is an estimate of the false alarm rate based on the
evaluation test results and the given test conditions.
You can calculate an upper confidence limit for P(fa) in the case of no mistakes. Let Ni be the
number of tests performed under the tight tank condition. Choose a confidence coefficient, (1 -
a), say 95 or 90 percent. Then the upper confidence limit, UL, for P(fa) is calculated as:
In the case of 0 false alarms out of 21 tests, the upper limit to P(fa) becomes 0.133 or 13.3
percent with a 95 percent confidence coefficient. That is, P(fa) is estimated at 0 percent, and
with a confidence of 95 percent, P(fa) is less than or equal to 13.3 percent. In general, you can
calculate the confidence interval for P(fa) from the binomial distribution with Ni trials. Calculate
and report the 95 percent confidence interval on the results form in Appendix C.
5.2.2 Probability Of Detecting A Leak, P(d)
Calculate the P(d) for a specific size of leak. Also report the size of leak that can be detected
with this probability. Normally this will be 0.10 gal/hr, as required by the performance
standards. The exception to this occurs if a method is tested using induced leak rates smaller
than 0.10 gal/hr, for example 0.05 gal/hr. Report the probability of detection, P(d), together with
the maximum leak rate used in the evaluation testing. The leak rate corresponding to the P(d) is
0.10 gal/hr or less.
Use the results obtained from the tests performed under induced leak conditions of leak rates less
than or equal to 0.10 gal/hr to calculate P(d). Let N2 be the number of such tests. Let TL2 be the
number of cases where the method indicated a leak. As before, the test results, Li are coded as 0
when the tank is declared tight and 1 when the tank is declared to be leaking. Thus, TL2 is
calculated as
where the sum is taken over the N2 tests with induced leaks. Estimate the P(d) by the ratio
1
UL for P(fa) = 1 — aNi
i=l
P(D) = TL2/N2
51

-------
The estimated P(d) must be at least 95 percent for the method to meet the performance standards.
Thus, TL2 must be either 20 or 21 out of 21 tests for the estimated probability of detection to be
at least 95 percent.
If the method identified the tank to be leaking in all tests where a leak was simulated, then the
proportion detected becomes 100 percent. However, this does not mean the method is perfect.
The P(d) of 100 percent is an estimate of the P(d), based on the evaluation test results and the
given test conditions.
You can calculate a lower confidence limit for P(d) in the case of no mistakes. Let N2 be the
number of tests performed under the induced leak conditions. Choose a confidence coefficient,
(1- a), for example, 95 or 90 percent. Then calculate the lower confidence limit, LL, for P(d) as:
LL for P(d) = ct1/^
In the case of correct identification of the 21 tests performed under leak conditions, the lower
limit to P(d) becomes 0.867 or 86.7 percent with a 95 percent confidence coefficient. P(d) is
estimated at 100 percent, and with a confidence of 95 percent, P(d) is greater than or equal to
86.7 percent. Calculate the 95 percent confidence interval for P(d) based on the binomial
distribution with N2 trials and reported on the results form in Appendix C.
5.3 Other Reported Calculations
This section describes other calculations needed to complete the Results Of U.S. EPA Standard
Evaluation form in Appendix C. Most of these calculations are straightforward and are
described here to provide complete instructions for the use of the results form.
These sections are only required if they are applicable to the particular non-volumetric method
being evaluated. If a section is not applicable or NA, skip the calculations and report NA on the
results form.
Size Of Tank
The evaluation results apply to volumes of tanks, interstitial spaces, and sumps up to 50 percent
larger capacity than the test volume for single tanks, interstices, sumps and other equipment, but
25 percent larger for tanks connected by siphon piping. The evaluation results also apply to all
smaller volumes for single tanks and tank systems with tanks connected by siphon piping.
Multiply the volume of the test volume by 1.5 for single tanks or other equipment and 1.25 for
tank systems connected by siphon piping. Round this number to the nearest 100 gallons and
report the result on page 2 of the results form. This allowance does not apply to all test methods
such as vacuum decay methods.
Maximum Allowable Temperature Difference
This section only applies if temperature conditioning was needed and used as part of the
evaluation procedure. If temperature does not affect the operation of the method, ignore this
section and indicate NA on the results form.
52

-------
Calculate the standard deviation of the temperature differences actually achieved during testing.
These are the differences between the product in the tank and the product added to fill the tank at
each fill. Multiply this number by the factor ±1.5 and report the result as the temperature range
on the limitations section of the results form. Notice that these temperature differences are
generally larger than in previous evaluations because the current protocol calls for a third of the
differences to be 10°F warmer than the product in the tank and a third of the differences to be
10°F cooler than the product in the tank. Previously the difference called for was ± 5°F. The ±
5°F difference was estimated to cover about 57 percent of the cases, while the ± 10°F range is
estimated to cover about 86 percent of the cases.2
Average Waiting Time After Filling
Calculate the average of the time intervals between the end of the filling cycle and the start of the
test for the 21 tests that started immediately after the specified waiting time. Note: If more than
21 tests are done immediately after the filling, use all such tests. However, do not use the time to
the start of the second test in a set, as this would give a misleading waiting time. On the results
form, report this average time as the waiting time after adding product. Note: You can use the
median as the average instead of the mean if there are atypical waiting times.
For tracer methods, the average waiting time may more appropriately be the time from adding
the tracer to the tank until completing the test.
Average Waiting After Topping Off
If the method fills the tank up into the fill pipe, calculate the average time interval between the
time when the final topping off was completed and the start of the test. Calculate this average
using data from all tests when this step was performed. Report the result on the results form as
the waiting time after topping off to the final testing level. If this step is not performed, for
example for a test with the tank at 95 percent of capacity, enter NA in the appropriate space on
the results form. Note: You can use the median instead of the mean if there are some atypical
waiting times.
Average Data Collection Time Per Test
Use the duration of the data collection phase of the tests to calculate the average data collection
time for the total number of at least 42 tests. Report this time as the average data collection time
per test.
Product Level
If all tests are done at the same product level, report that level on the results form. If testing was
done at different levels, report the applicable product level as the acceptable range, for example
from 60 to 90 percent full, used in the testing.
2 "Typical Tank Testing Conditions," Flora, Jairus D., Jr., and Jean Pelkey, Report on Work Assignment 22, Task
13, EPA Contract No. 68-01-7383, Midwest Research Institute, December 2, 1988.
53

-------
Minimum Total Testing Time
Finally, calculate an average total test time from the test data. This is the time it takes from the
time the test crew arrives at the site until a test is completed, the method dismantled, and the tank
returned to service. Typically, it is the time from initial setup of the method through the first test
data collection, plus the time required to dismantle the equipment. Report this total time lapse on
the results form as the minimum time the tank is expected to be out of service for a test of this
type.
The intent of this is to provide an estimate of the time the testing requires. Testing generally
means that a tank must be taken out of service with no dispensing or delivery during the duration
of the test. Non-volumetric methods differ in those parts of their operation that require the tank
to be out of service. Report the estimated testing time this method requires.
5.4 Supplemental Data Analyses (Optional)
This section discusses some additional data analyses that may be possible with the data,
depending on the actual results. It also provides some rationale for the sample size selection.
One-Sided Confidence Limits On P(fa) And P(d)
It is possible to estimate the P(fa) and P(d) directly as done in Section 5.1 with any sample size.
However, for fewer than 20 tests, the estimate of P(fa) is 0 or exceeds 5 percent, depending on
whether any false alarms are found. Similarly, P(d) is 100 percent or less than 95 percent for
sample sizes less than 20, depending on whether any leaks are missed or not. Thus, the sample
size of 20 is the smallest that allows for one mistake in each case and still provides estimated
performance meeting the EPA standards. The sample size of 21 was chosen from test design
considerations to balance the different conditions.
Calculate confidence limits for P(fa) and P(d) based on the observed results and sample sizes.
The formulas for perfect scores were given in Section 5.2.1 for P(fa) and in Section 5.2.2 for
P(d). These also depend on the selected confidence coefficient. Table 6 below gives 90 and 95
percent one-sided confidence limits for P(fa) and P(d) based on samples of 21 tests for the case
of no mistakes and one mistake, the two conditions under which the method meets the EPA
performance standards, if evaluated with the minimum 21 tests.
Table 6. One Sided Confidence Limits For P(fa) And P(d)
Field Test
Confidence Coefficient
Results
90%
95%
0 Error out of 21
P(fa)< 0.104
P(fa)< 0.133
1 Error out of 21
P(fa)< 0.173
P(fa) < 0.207
0 Error out of 21
P(d) > 0.896
P(d) > 0.867
1 Error out of 21
P(d) > 0.827
P(d) > 0.793
54

-------
Table 6 shows the confidence limits start to become large for high confidence with even one
error. Using a larger sample size improves the confidence limits, but adds significantly to the
cost of testing. We selected the sample sizes as a compromise to provide reasonable estimates
while not requiring excessively expensive testing.
5.5 Sensor Performance Calculations
From the results obtained after completing testing, perform a series of calculations to evaluate
the sensor's performance. The results obtained from individual sensors do not fit the standard
P(fa) and P(d) results that are calculated for most of the release detection methods. The goal of
the testing is to determine the capabilities of each sensor to produce the correct sensor output
depending on the test. Each sensor has different capabilities and, therefore, will have different
data outputs. Table 7 presents the performance parameters and evaluation metrics are the means
of determining the operability of each sensor.
Table 7. Performance Parameters
Performance Parameter
Evaluation Metric
Data To Be Recorded
Average detection time
Average and SD of the difference in
actuation time and test start times
Test start time and actuation time
calculated for each test condition
Liquid activation height
(liquid only)
Average and SD of the activation height
Liquid height level at activation,
calculated for each liquid
Specificity
% Specificity
Detection data calculated for each test
condition by product
Accuracy
(qualitative only)
Relative % accuracy
Detection data calculated for each test
condition
Accuracy
(quantitative only)
% Accuracy
Detection data calculated for each test
condition
Precision
(quantitative only)
% Coefficient of variation
Detection data calculated for each test
condition
Average recovery time
Average and SD of the difference
between recovery and test end times
Test end time and recovery time
calculated for each test condition
Average Detection Time
Evaluate detection time for all sensors. Report the average detection time as the average (x) and
the standard deviation (SD) of the observed values for each repeated test condition.
Average Liquid Activation Height
Report the liquid activation height as the average and the SD of the observed values for each
repeated test condition.
55

-------
Table 8. Notation Summary For Water Sensor Readings At The jth Replicate
Increment No.
Calculated Level
Change
(inch)
A
Sensor Reading
(inch)
B
Measured
Sensor
Increment
(inch)
C
Increment
Difference
C alculated-Meas.
(inch)
C-A
1
+ h
Wi,j
Wi,j-Xj*
di.i
2
+ h
W2,,
W2,j-Wl,j
d2j
3
+ h
Wij
\V= - w2j
d3.
iij
+ h
w ¦
nj>J
W ¦ - W , ¦
nj,j nj —l,j
dn(,j
* Xj is the water level in inches detected for the first time by the sensor during the jth replication of the test.
Specificity
Calculate the percent or % specificity using the following equation for each liquid individually as
follows:
Specificity, % = 100 x i
x = mean of observed values, cm
xt= the theoretical value, cm
Accuracy For Qualitative Sensors Only
Determine the accuracy for the qualitative liquid and vapor sensors by calculating the percent
accuracy of replicates of the tested products individually as follows:
Accuracy, %= 100 X
r= the number of positive responses
n = the number of tests for a particular liquid or test gas
Relative Percent Accuracy For Quantitative Sensors Only
Use the following equation to compute the accuracy in measuring the liquid level for each
replicate measured of the tested product:
\M — D\
Accuracy, % =	—— x 100
56

-------
M = measured liquid level, cm, or established gas concentration, ppmv
D = detected liquid level, cm, or detected gas concentration, ppmv
Similarly, compute the vapor sensors relative percent accuracy for each set of replicates under a
test condition.
Precision For Quantitative Sensors Only
Calculate precision as the percent coefficient of variation, or %CV, for quantitative liquid and
vapor sensors as follows:
SD = standard deviation of n values, cm or ppmv
x = mean of observed values, cm or ppmv
Average Recovery Time
For liquid sensors, the recovery time is how long it takes the sensor to return to an inactivated
state after it is removed from the testing condition. Record, average, and report this time. Some
sensors are instantaneous and can be reported as such. Others take time to be prepared for the
next test condition.
For vapor sensors, the recovery time is dependent on whether a sensor gives a quantitative or
qualitative response. The recovery time for a quantitative sensor is when the output returns to
within 5 percent of the original stable baseline level. Calculate the 5 percent stable baseline level
according to the following equation.
BL = stable baseline output, ppmv
HL = stable high level output, ppmv
The recovery time for a qualitative output sensor is defined as when the sensor goes from
activated state to an inactivated state.
5% Stable baseline output, ppmv = BL + (HL — BL) x 0.05
57

-------
Section 6: Interpretation
The results reported are valid for the test design conditions during the evaluation, which have
been chosen to represent situations commonly encountered in the field. These should be typical
of most tank testing conditions, but extreme conditions can occur and might adversely affect the
performance of the method. It is emphasized that the performance estimates are based on
average results obtained in the tests. An individual test may not do as well. Some individual
tests may do better.
6.1	Basic Performance Estimates
The relevant performance measures for proving that a TTT method meets EPA standards are the
P(fa) and P(d) for a leak rate of 0.10 gal/hr. Compare the estimated P(fa) with EPA's standard of
P(fa) not to exceed 5 percent. In general, a lower P(fa) is preferable, since it implies the chance
of mistakenly indicating a leak on a tight tank is less. A general goal is to reduce the number of
false alarms. However, reducing the false alarm rate may also reduce the chance of detecting a
leak. The probability of detection generally increases with the size of the leak. EPA's standard
specifies that P(d) be at least 95 percent for a leak of 0.10 gal/hr. A higher estimated P(d) means
there is less chance of missing a small leak.
The discrete nature of the data implies that only a few values of P(fa) or P(d) are possible. With
the standard 21 tests for each test condition of a tight or leaking tank, the possible values are 0,
1/21, 2/21, etc. Consequently, the reported estimates are only precise to about 5 percent. The
confidence limits reported in the case of a perfect score indicate the expected range of the true
P(fa) or P(d). For example, a method that achieved 0 false alarms throughout testing would not
be expected to have a 0 false alarm rate. Instead, its false alarm rate should be less than 10.4
percent with 95 percent confidence.
If testing is done at an induced leak rate less than 0.10 gal/hr, the P(d) may be reported at the
smaller leak rate actually used. The standard test, using an induced leak rate of 0.10 gal/hr,
would report P(d) for the rate of 0.10 gal/hr. In general, a method that can detect a smaller leak
with high probability is preferred because it identifies a potential problem earlier. This may
reduce the amount of pollution and the cost of remedial action.
6.2	Limitations
Report the limitations on the evaluation results section of the Results Of U.S. EPA Standard
Evaluation form. The intent is to document that the results are valid under conditions
represented by the test conditions. The test conditions were chosen to represent the majority of
testing situations, but do not include the most extreme conditions under which testing could be
done. The test conditions were also selected to be practical and not impose an undue burden for
evaluation on the evaluator.
For volumetric methods, one limitation of the results is the size of the tank. Tests based on
volumetric changes generally perform less well as the size of the tank increases. Consequently,
you may apply the results of the evaluation to tanks smaller than the test tank. The results may
58

-------
also be extended to volumes of tanks, interstitial spaces, and sump of 50 percent larger capacity
than the test volume for single tanks or other equipment and 25 percent larger capacity for tank
systems connected by siphon piping.
Optional testing on tanks connected by siphon piping evaluates the method's performance in
different tank configurations. The requirement for an evaluation on tanks connected by siphon
piping is to perform 24 tests for volumetric methods or 42 tests for non-volumetric methods. If
the estimated probabilities are within the regulatory limit with the additional tests, you may use
the method in siphoned systems. The results are limited to one more tank in the siphoned
configuration than used in the evaluation testing.
Another limitation on the results is the temperature differential between the product added to the
tank and that of the product already in the tank. The temperature differential is a factor when
performing a test shortly after a tank is filled. The reported results are applicable provided the
temperature differential is no more than that used in the evaluation. The results cannot be
guaranteed for temperature differentials larger than those used in the evaluation.
Non-volumetric TTT methods based on different operating principles have different factors that
can interfere with their performance. Consequently, the limitations on the applicability of the
performance estimates also vary with the method. However, there may be interfering factors
other than those listed in the test design that affect a particular test method. If so, those
additional factors might limit the applicability of the method. The reporting form provides a
place to identify other sources of interference and to state the test conditions for them.
Some non-volumetric test methods use more than one mode of operation. If so, different
limitations may apply to each mode of release detection. It is possible that one mode of
operation may be unaffected by size of tank, but that another may depend strongly on tank size.
For example, a water sensor may be used to test for leaks in the presence of a high groundwater
level. It may do so by sensing water incursion, in which case it must be able to detect water
incursion at the rate of 0.10 gal/hr. Since the time required for the water level to be detectable at
a fixed rate of incursion will be a function of the size of the tank, this mode of release detection
is dependent on tank size.
6.3 Additional Calculations
If the performance estimates do not meet the performance requirements, the vendor may want to
investigate the conditions under which errors occurred. Calculating the percent of errors by size
of leak, temperature condition, and length of stabilization time may suggest ways to improve the
method. This may be as straightforward as identifying conditions that lead to poor performance
and revising the operating procedure to avoid those, or it may require redesign of the method.
The relationship of performance to test conditions is primarily of interest when the method does
not meet EPA's performance standards. Developing these relationships is part of the optional or
supplementary data analysis that may be useful to the vendor, but not to many tank owners or
operators.
59

-------
Section 7: Reporting Of Results
Appendices B and C are designed to be the framework for a standard report for volumetric and
non-volumetric TTT methods, respectively. There are four parts to the results report, each with
instructions for completion.
•	Results Of U.S. EPA Standard Evaluation form - This form is an executive summary of
the findings and is for tank owners or operators who use this method of release detection.
The results form is easy to reproduce for wide distribution.
•	Description (volumetric or non-volumetric) tank tightness method - The evaluator,
assisted by the vendor, completes the description form.
•	Reporting form for leak rate data - This table summarizes the test results and contains the
information on starting dates and times, test duration, and leak test results.
•	Individual test log - While the individual test log is designed to be flexible, you may
need to modify it for some test methods. Use this to record data in the field. The
evaluator must maintain test logs but they are not mandatory for the standard report.
These serve as the backup data to document the performance estimates reported.
A method that uses more than one mode of release detection may achieve different performance
results for the different modes of operation. The results form is structured to allow for reporting
the P(fa) and P(d) separately for different modes of release detection. The method meets EPA's
performance requirements only if all modes of release detection meet those requirements. The
statement of compliance with EPA's performance standards must be consistent with stated
limitations on the form and with the standard operation of the method as described on the
description form.
Suppose that a method has two modes of testing, a basic one and an ancillary one for testing in
the presence of a high groundwater level. Suppose the test method when evaluated in the case of
high groundwater level does not meet EPA's performance requirements, but the basic one does.
Then you can issue a report, stating the method meets EPA's performance requirements, but
cannot test when the groundwater level is above the bottom of the tank.
Non-volumetric methods may require some modification of the forms. If the forms need to be
modified, the evaluator makes the required modifications and uses the resulting forms. Record
the conditions during the evaluation tests and the factors that affect the performance of the
method. Test conditions actually used and reported may limit the performance results.
60

-------
Appendix A
Definitions And Student's t Distribution
A-l

-------
Definitions of terms used throughout the test procedures and the Student's t distribution table in
Table A-l are presented here. For more information on the statistical approach and relationships
between the statistics calculated in these test procedures see General Guidance For Usins EPA 's
Standard Test Procedures For Evaluating Release Detection Methods.
Calculated Leak Rate, R A positive number, in gallons per hour (gal/hr), estimated
by the TTT method and indicating the amount of product
leaking out of the tank. A negative leak rate could result
from water leaking into the tank, miscalibration, or other
causes.
The actual leak rate, in gal/hr, introduced in the evaluation
data sets, against which the results from a given method
will be compared.
The leak rate above which a method declares a leak. It is
also called the threshold of the method.
Declaring that a tank is leaking when in fact it is tight.
The probability of declaring a tank leaking when it is tight.
In statistical terms, this is also called the Type I error, and
is denoted by alpha (a). It is usually expressed in percent,
as 5 percent.
The probability of detecting a leak rate of a given size, R
gal/hr. In statistical terms, it is the power of the test
method and is calculated as one minus beta (P), where beta
is the probability of not detecting or missing a leak rate R.
Commonly the power of a test is expressed in percent, as
95 percent.
The average difference between calculated and induced
leak rates. It is an indication of whether the TTT method
consistently overestimates as a positive bias or
underestimates as a negative bias the actual leak rate.
Mean Squared Error, MSE An estimate of the overall performance of a test method.
Induced Leak Rate, S
Threshold, Th
False Alarm
Probability Of False Alarm,
P(fa)
Probability Of Detection,
P(d)
Method Bias, B
Root Mean Squared Error,
RMSE
Precision
The positive square root of the mean squared error.
A measure of the test method's ability in producing similar
results, that is, in close agreement, under identical
conditions. Statistically, the precision is expressed as the
standard deviation of these measurements.
A-2

-------
Accuracy
Variance:
df
a = .10
a = .05
a = .025
a = .010
a = .005
1
3.078
6.314
12.706
31.821
63.657
2
1.886
2.920
4.303
6.965
9.925
3
1.638
2.353
3.182
4.541
5.841
4
1.333
2.132
2.776
3.747
4.604
5
1.476
2.015
2.571
3.365
4.032
6
1.440
1.943
2.447
3.143
3.707
7
1.415
1.895
2.365
2.998
3.499
8
1.397
1.860
2.306
2.896
3.355
9
1.383
1.833
2.262
2.821
3.250
10
1.372
1.812
2.228
2.764
3.169
11
1.363
1.796
2.201
2.718
3.106
12
1.356
1.782
2.179
2.681
3.055
13
1.350
1.771
2.160
2.650
3.012
14
1.345
1.761
2.145
2.624
2.977
15
1.341
1.753
2.131
2.602
2.947
16
1.337
1.746
2.120
2.583
2.921
17
1.333
1.740
2.110
2.567
2.898
18
1.330
1.734
2.101
2.552
2.878
19
1.328
1.729
2.093
2.539
2.861
20
1.325
1.725
2.086
2.528
2.845
21
1.323
1.721
2.080
2.518
2.831






22
1.321
1.717
2.074
2.508
2.819
23
1.319
1.714
2.069
2.500
2.807
24
1.318
1.711
2.064
2.492
2.797
A-3
The degree to which the calculated leak rate agrees with
the induced leak rate on the average. If a method is
accurate, it has a very small or 0 bias.
A measure of the variability of measurements. It is the
square of the standard deviation.
Table A-l. Percentage Points Of Student's t Distribution
f(t)
o
t

-------
df
a = .10
a = .05
a = .025
a = .010
a = .005
25
1.316
1.708
2.060
2.485
2.787
26
1.315
1.706
2.056
2.479
2.779
27
1.314
1.703
2.052
2.473
2.771
28
1.313
1.701
2.048
2.467
2.763
29
1.311
1.699
2.045
2.462
2.756
30
1.310
1.697
2.042
2.457
2.750
40
1.303
1.684
2.021
2.423
2.704
60
1.296
1.671
2.000
2.390
2.660
120
1.289
1.658
1.980
2.358
2.617
inf.
1.282
1.645
1.960
2.326
2.576
A-4

-------
Appendix B
Volumetric Methods Reporting Forms
B-l

-------
Volumetric Method Evaluation Forms
Appendix B provides four sets of blank forms. Once completed, these forms provide the
framework for a standard report. They consist of:
1.	Results Of U.S. EPA Standard Evaluation - Volumetric Tank Tightness Testing
Method
2.	Description Of Volumetric Tank Tightness Testing Method
3.	Reporting Form For Leak Rate Data - Volumetric Tank Tightness Testing Method
4.	Individual Test Logs - Volumetric Tank Tightness Testing Method
Each set of forms includes instructions on how to complete the forms and who should complete
them. The following is an overview on various responsibilities.
1.	Results Of U.S. EPA Standard Evaluation - The evaluator completes this form at the
end of the evaluation.
2.	Description Of Volumetric Tank Tightness Testing Method - The evaluator assisted
by the vendor or a field crew completes this form at the end of the evaluation.
3.	Reporting Form For Leak Rate Data - The evaluator or statistician analyzing the data
completes this form. You can develop a blank form on a personal computer, generate
the database for a given evaluation, and merge the two on the computer. You can
also complete this form manually. The input for the form consists of the field test
results recorded by the evaluator's field crew on the individual test logs discussed
below and the vendor's test results.
4.	Individual Test Logs - The evaluator completes these forms. Keep these forms blind
to the vendor's field crew. Reproduce a sufficient number of at least 24 copies of the
blank form provided in this appendix and produce a bound notebook for the complete
test period.
After completing the evaluation, the evaluator collates all the forms into a single standard report
in the order listed above.
Distribution Of The Evaluation Test Results
The evaluator performing the evaluation prepares a report for the vendor describing the results of
the evaluation. The evaluator of the release detection method provides the report to the vendor.
The vendor is responsible for distributing the results to tank owners or operators and to
regulators.
This report consists primarily of the forms in this appendix. The first form reports the results of
the evaluation. This two-page form is designed to be distributed widely. Provide a copy of this
two-page form to each tank owner or operator who uses this method of release detection. The
owner or operator must retain a copy of this form as part of his recordkeeping requirements. The
owner or operator must also retain copies of each tank test performed at his facility to document
the tanks passed the tightness test. The vendor distributes this two-page form to regulators who
must approve release detection methods for use in their jurisdiction.
B-2

-------
The evaluator submits the completed report, consisting of all the forms in Appendix B, to the
release detection method vendor. The vendor may distribute the complete report to regulators
who wish to see the data collected during the evaluation. The vendor may also distribute the
report to release detection method customers who want to see additional information before
deciding to select a particular release detection method.
The evaluator reports to the vendor any optional calculations made regarding the release
detection method. The vendor may use this report to understand the details of the performance
and perhaps improve the method. The vendor has the discretion to distribute this form.
B-3

-------
Results Of U.S. EPA Standard Evaluation
Volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator must fill out this form after completing the evaluation of the method. This form
contains the most important information relative to the method evaluation. Fill out all items and
check the appropriate boxes. If a question is not applicable to the method, write NA in the
appropriate space.
This form consists of five main parts:
1.	Method Description
2.	Evaluation Results
3.	Test Conditions
4.	Limitations On The Results
5.	Certification Of Results
Method Description
Indicate the commercial name of the method, the version, and the name, address, and telephone
number of the vendor. Some vendors use different versions of their method when using it with
different products or tank sizes. If so, indicate the version used in the evaluation. If the vendor
is not the party responsible for developing and using the method, then indicate the home office
name and address of the responsible party.
Evaluation Results
The vendor supplies the method's Th (threshold). This is the criterion for declaring a tank to be
leaking. Typically, a method declares a tank to be leaking if the measured leak rate exceeds Th.
P(fa) is the probability of false alarm calculated. Report P(fa) in percent rounded to the nearest
whole percent.
P(d) is the probability of detecting a leak rate of 0.10 gallon per hour (gal/hr) and is calculated.
Report P(d) in percent rounded to the nearest whole percent.
If the P(fa) calculated is 5 percent or less and if the P(d) calculated is 95 percent or more, then
check the does box. Otherwise, check the does not box.
Test Conditions During Evaluation
Insert the information in the blanks provided. Request the nominal volume of the tank in gallons
and the tank material, for example, steel or fiberglass. Also, give the tank diameter and length in
inches. Report the product used during the testing. If a level lower than a 90-95 percent full
level is used, justify the use and note that the method is limited to testing a tank below the liquid
B-4

-------
level used. This also restricts use of tanks tested by this method to no more than the level
for which the method is approved. Give the range of temperature differences actually
measured, as well as the standard deviation of the observed temperature differences. Also,
indicate the level in the tank at which the testing was done.
Limitations On The Results
The size in gallons of the largest tank to which these results can be applied is calculated as 1.5
times the size in gallons of the test tank. This allowance does not apply to all test methods such
as vacuum decay methods. For tank systems with tanks connected by siphon piping, the results
are limited to the number of tanks in the manifold used in testing. The volume limit applies to
the total volume of the tank system with tanks connected by siphon piping.
If the method compensates for groundwater levels above the bottom of a tank, then check the can
box. Otherwise, check the cannot box.
Certification Of Results
The evaluator provides his or her name and signature, and the name, address, and telephone
number of the organization.
B-5

-------
Results Of U.S. EPA Standard Evaluation
Volumetric Tank Tightness Testing Method
This form tells whether the tank tightness testing method described below complies with the
performance requirements of the federal underground storage tank regulation. The evaluation
was conducted by the vendor according to the U.S. EPA's Standard Test Procedure for
Evaluating Release Detection Methods: Volumetric Tank Tightness Testing Methods. The full
evaluation report also includes a form describing the method and a form summarizing the test
data.
Tank owners using this release detection system should keep this form on file to prove
compliance with the federal UST regulation. Tank owners should check with regulatory
authorities to make sure this form satisfies their requirements.
1. Method Description
Method name
Version
number
Vendor
Street address
City, state, zip
Telephone number
Protection Agency of 0.10 gallon per hour at
P(d) of 95% and P(fa) of 5%.
3. Test Conditions During Evaluation
a. The evaluation testing was conducted in a
gallon
steel
~ tank ~ fiberglass tank
inches in diameter
inches in length
2. Evaluation Results
b. The tests were conducted with the tank
	% full.
a.	This method, which declares a tank to be
leaking when the measured leak rate exceeds
the threshold of	gallon per hour,
has a probability of false alarms [P(fa)] of
0/
	/O.
b.	The corresponding probability of detection
[P(d)] of a 0.10 gallon per hour leak is	
0/
	/O.
c.	Therefore, this method
~ does ~ does not
meet the federal performance standards
established by the U.S. Environmental
c. The temperature difference between product
added to fill the tank and product already in
the tank ranged from
	 °F to 	
with a standard deviation
of:	°F
d. The product used in the evaluation was
4. Limitations On The Results
•	The method has not been substantially
changed.
•	The vendor's instructions for using the
method are followed.
Volumetric TTT Method - Results Form
Page 1 of 2

-------
Results Of U.S. EPA Standard Evaluation
Volumetric Tank Tightness Testing Method
The tank is no larger than	
gallons.
The tank contains a product identified on the
method description form.
The waiting time after adding any substantial
amount of product to the tank is at least
	hours.
The temperature of the added product does
not differ more than	
degrees Fahrenheit from already in the tank.
The waiting time between the end of topping
off, if any, and the start of the test data
collection is at least	hours.
The total data collection time for the test is
at least	hours.
•	This method can be used on up to	of
tanks connected by siphon piping with a
total volume of	.
•	This method
~ can ~ cannot
be used if the groundwater level is above the
bottom of the tank.
•	Other limitations specified by the vendor or
determined during testing:
Safety disclaimer: This test procedure only
addresses the issue of the method's ability to
detect leaks. It does not test the method for
safety hazards.
Additional explanations or comments
5. Certification Of Results
I certify that the tank tightness test method was operated according to the vendor's instructions. I also
certify that I performed the evaluation according to the procedure specified by EPA and that the
results presented in the report were obtained during the evaluation.
Printed name
Organization performing evaluation
Signature
Street address
Date	City, state, zip
Phone number
Volumetric TTT Method - Results Form
Page 2 of 2

-------
Description Of Volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator, with assistance from the vendor, must fill out this form upon completion of the
evaluation of the method. This form provides supporting information on the principles behind
the method or on how the method works.
To minimize the time to complete this form, we provide the most frequently asked questions.
For answers that depend on site conditions, give answers that apply in typical conditions. Write
in any additional information about the testing method you believe is important.
There are seven parts to this form:
1.	Method Name and Version
2.	Product Description
•	Product type
•	Product level
3.	Level Measurement
4.	Temperature Measurement
5.	Data Acquisition
6.	Procedure Information
•	Waiting times
•	Test duration
•	Total time
•	Identifying and correcting for interfering factors
•	Interpreting test results
7.	Exceptions
Indicate the commercial name and the version of the method in the first part.
Note: The version is provided for methods that use different versions of the method for different
products or tank sizes.
For the six remaining parts, check all appropriate boxes for each question. Check more than one
box per question if it applies. If you check the other box, use the space provided to specify or
briefly describe the matter. If necessary, use the white space next to a question for a description.
B-6

-------
Description
Volumetric Tank Tightness Testing Method
1. Method Description
Method name 	
Version name
2. Product Description
Product type
a. For what products can this method be used? Check all applicable.
~	Gasoline	~	Diesel
~	Aviation fuel ~	Fuel oil #4
~	Fuel oil #6	~	Solvent
~	Waste oil	~	Other, list
Product level
b. What minimum product level is required to conduct a test?
~ Above grade ~ Within the fill pipe
I I Greater than 90% full Q Greater than 50% full
I I Other, specify		
c.	Is a method used to add or withdraw product to maintain a constant level of product?
I I Yes HH No
d.	Does the method measure inflow of water as well as loss of product at gallons per hour?
I I Yes O No
e.	Does the method detect the presence of water in the bottom of the tank?
I I Yes O No
3. Level Measurement
a. What technique is used to measure changes in product volume?
~	Directly measure the volume of product change ~ Changes in head
pressure
~	Changes in buoyancy of a probe	~ Mechanical level
measure; for
example, ruler,
dipstick
~	Changes in capacitance	~ Ultrasonic
~	Change in level of float; specify principle, for example, capacitance,
magnetostrictive, and load cell
Volumetric TTT Method - Description
Page 1 of 7

-------
Description
Volumetric Tank Tightness Testing Method
~ Other, describe briefly
4. Temperature Measurement
a.	If product temperature is measured during a test, how many temperature sensors are used?
~	Single sensor, without circulation ~ Single sensor, with circulation
~	2-4 sensors	~ 5 or more sensors
~	Temperature-averaging probe
b.	If product temperature is measured during a test, what type of temperature sensor is used?
~	Resistance temperature Q Bimetallic
detector (RTD)	strip
~	Quartz crystal	~ Thermistor
~	Other, describe
c. If product temperature is not measured during a test, why not?
I I The factor measured for change in level or
volume is independent of temperature, for
example, mass
I I The factor measured for change in level or
volume self-compensates for changes in
temperature
I I Other, explain briefly
5. Data Acquisition
a. How are the test data acquired and recorded?
~	Manually ~ By strip chart
~	By computer
Volumetric TTT Method - Description
Page 2 of 7

-------
Description
Volumetric Tank Tightness Testing Method
6. Procedure Information
Waiting times
a. What is the minimum waiting period between adding a large volume of product to bring the level to test
requirements and the beginning of the test, for example, from 50 percent to 95 percent capacity?
~	No waiting period	~ < 3 hours
~	3-6 hours	~ 7-12 hours
I I >12 hours
~	Variable, depending on tank size, amount added, operator discretion
b. What is the minimum waiting period between topping off the tank by adding a small amount of product to fine
tune the desired level for testing, for example, from 2 inches to 5 inches above grade and beginning the test?
~	No waiting period ~ < 1 hours
~	1-2 hours	~ > 2 hours
~	Variable, depending on the amount of
product added
Test duration
c. What is the minimum time for collecting data?
~	< 1 hour	~	1 hour
~	2 hours	HH	3 hours
~	4 hours	HH	5-10 hours
~	More than 10 hours	~	Variable
Total time
d. What is the total time needed to test with this method? This includes setup time plus waiting time plus testing
time plus time to return tank to service.
Hours	Minutes
e. What is the sampling frequency for the level and temperature measurements?
~	More than once per second ~
~	Every 1-15 minutes	~
~	Every 31-60 minutes	~
~	Variable
At least once per minute
Every 16-30 minutes
Less than once per hour
Identifying and correcting for interfering factors
f. How does the method determine the presence and level of the groundwater above the bottom of the tank?
~	Observation well near tank
I I Information from USGS
~	Information from personnel on-site
Volumetric TTT Method - Description	Page 3 of 7

-------
Description
Volumetric Tank Tightness Testing Method
~	Presence of water in the tank
~	Level of groundwater above bottom of the
tank not determined
~	Other, describe briefly
g. How does the method correct for the interference due to the presence of groundwater above the bottom of the
tank?
~	Head pressure increased by raising the level
of the product
~	Different head pressures tested and leak
rates compared
~	Method tests for changes in water level in
tank
~	No action
~	Other, describe briefly
h. How does the method identify the presence of vapor pockets?
~	Erratic temperature, level, or temperature-
compensated volume readings
~	Sudden large changes in readings
~	Statistical analysis of variability of readings
~	Not applicable; underfilled test method used
~	Not identified
~	Other, describe briefly
i. How does the method correct for the presence of vapor pockets?
~	Bleed off vapor and start test over
~	Identify periods of pocket movement and
discount data from analysis
~	Not corrected
~	Not applicable; underfilled test method
used
I I Other, describe briefly
j. Are the temperature and level sensors calibrated before each test?
I I Yes O No
k. If not, how often are the sensors calibrated?
I I Weekly	O Monthly
~ Yearly or less frequently ~ Never
Volumetric TTT Method - Description	Page 4 of 7

-------
Description
Volumetric Tank Tightness Testing Method
Interpreting test results
1. How are level changes converted to volume changes; that is, how is height-to-volume conversion factor
determined?
~	Actual level changes observed when known
volume is added or removed, for example,
liquid, metal bar
I I Theoretical ratio calculated from tank
geometry
I I Interpolation from tank vendor's chart
I I Not applicable; volume measured directly
~	Other, describe briefly
m. How is the coefficient of thermal expansion (Ce) of the product determined?
I I Product sample taken for each test and Ce
determined from specific gravity
I I Value supplied by vendor of product
I I Average value for type of product
I I Other, describe briefly
n. How is the leak rate of gallon per hour calculated?
I I Average of subsets of all data collected
I I Difference between first and last data collected
I I From data to last	hours of test period
I I From data determined valid by statistical
analysis
~ Other, describe briefly
o. What threshold value for product volume change of gallon per hour is used to declare a tank is leaking?
I I 0.05 gallon/hour Q 0.10 gallon/hour
~	0.20 gallon /hour
~	Other, describe 	
p. Under what conditions are test results considered inconclusive?
I I Groundwater level above bottom of tank
I I Presence of vapor pockets
I I Too much variability in the data of standard
deviation beyond a given value
~ Unexplained product volume increase
Volumetric TTT Method - Description	Page 5 of 7

-------
Description
Volumetric Tank Tightness Testing Method
I I Other, describe briefly
7. Exceptions
a. What are acceptable deviations from the standard testing test procedures?
~	None	~ Length the duration of
test
~	Other, describe 	
b. What are the conditions under which a test should not be conducted?
I I Groundwater level above bottom of tank
I I Presence of vapor pockets
I I Large difference between ground temperature
and delivered product temperature
~ High ambient temperature
I I Invalid for some products, specify
I I Other, describe briefly
c. What elements of the test procedure are determined by testing personnel on-site?
~	Waiting period between filling tank and
beginning test
~	Length of test
~	Determination of presence of vapor pockets
I I None
~	Other, describe briefly
Volumetric TTT Method - Description
Page 6 of 7

-------
Description
Volumetric Tank Tightness Testing Method
Additional explanations or comments
Volumetric TTT Method - Description
Page 7 of 7

-------
Reporting Form For Leak Rate Data
Volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator must fill out this form upon completion of the evaluation of the method. A single
sheet provides for 24 test results, the minimum number of tests required in the test procedures.
Use as many pages as necessary to summarize all of the tests attempted.
Indicate the commercial name and the version of the method and the period of evaluation above
the table. You may use different versions of the method for different products or tank sizes.
The evaluator or the statistician analyzing the data completes this form. Develop a blank form
on a personal computer, generate the database for a given evaluation, and merge the two on the
computer. You can complete the form manually. The input for the form consists of the field test
results recorded by the evaluator's field crew on the individual test logs and the vendor's test
results.
The table consists of 11 columns. One line is provided for each test performed during evaluation
of the method. If a test was invalid or aborted, list the test with the appropriate notation, such as
invalid, on the line.
The test number in the first column refers to the test number from the randomization design
determined according to the instructions in Section 6 of the test procedures. Since some changes
to the design might occur during the course of field-testing, the test numbers might not always be
in sequential order.
Note: Report the results from the trial run here as well.
The following list matches the column input required with its source, for each column in the
table.
Column No.	Input	Source
1
Test number or trial run
Randomization design
2
Date at completion of last fill
Individual test log
3
Time at completion of last fill
Individual test log
4
Date test began
Individual test log
5
Time test began
Individual test log
6
Time test ended
Individual test log
7
Product temperature differential
Individual test log
8
Nominal leak rate
Randomization design
9
Induced leak rate
Individual test log
10
Measured leak rate
Vendor's records
11
Measured minus induced leak rate
By subtraction
B-7

-------
Reporting Form For Leak Rate Data
Volumetric Tank Tightness Testing Method
Method name and version
Evaluation period from	to	(dates)

Date At
Time At



Product





Completion
Completion
Date Test
Time Test
Time Test
Temperature
Nominal
Induced
Measured
Meas.-Ind.
Test
Of Last Fill
Of Last Fill
Began
Began
Ended
Differential
Leak Rate
Leak Rate
Leak Rate
Leak Rate
No.
(m/d/y)
(military)
(m/d/y)
(military)
(military)
(°F)
(gal/hr)
(gal/hr)
(gal/hr)
(gal/hr)
Trial





0
0
0


Run


















1










2










3










4










5










6










7










8










9










10










11










12










13










14










15










16










17










18










19










20










21










22










23










24










Volumetric TTT Method-Reporting Form
Page 1 of 2

-------
Reporting Form For Leak Rate Data
Volumetric Tank Tightness Testing Method
Method name and version
Evaluation period from	to	(dates)

Date At
Time At



Product





Completion
Completion
Date Test
Time Test
Time Test
Temperature
Nominal
Induced
Measured
Meas.-Ind.
Test
Of Last Fill
Of Last Fill
Began
Began
Ended
Differential
Leak Rate
Leak Rate
Leak Rate
Leak Rate
No.
(m/d/y)
(military)
(m/d/y)
(military)
(military)
(°F)
(gal/hr)
(gal/hr)
(gal/hr)
(gal/hr)
25










26










27










28










29










30










31










32










33










34










35










36










37










38










39










40










41










42










43










44










45










46










47










48










Volumetric TTT Method-Reporting Form
Page 2 of 2

-------
Individual Test Log
Volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator's field crew fills out this form. Complete a separate form for at least 24 individual
tests. Keep the information on these forms blind to the vendor during the period of evaluation of
their method.
The form consists of nine parts:
1.	Header information
2.	General background information
3.	Conditions before testing
4.	Topping off records, if applicable
5.	Conditions at beginning of test
6.	Conditions at completion of testing
7.	Leak rate data
8.	Additional comments, if needed
9.	Induced leak rate data sheets
Fill out all items and check the appropriate boxes. If a question is not applicable, then indicate as
NA. The following provides guidance on the use of this form.
Header Information
Repeat the header information on all five pages, if used. If a page is not used, cross it out and
initial it. The evaluator's field operator needs to print and sign his or her name and note the date
of the test on top of each sheet.
The test number is the number obtained from the randomization design. It is not the sequential
running test number. If you repeat a test, indicate the test number on the test log, for example,
test no. 5 repeat.
General Background Information
Indicate the commercial name of the method. Include a version identification if the method uses
different versions for different products or tank sizes. Prior to testing, obtain the vendor's
recommended stabilization period. This is important since it will influence scheduling of the
evaluation. All other items in this section refer to the test tank and product. Indicate the
groundwater level at the time of the test.
Theoretically, this information remains unchanged for the whole evaluation period. However,
weather conditions could change and affect the groundwater level, or the evaluator could change
the test tank.
B-8

-------
Conditions Before Testing
Fill in all the blanks. If obtaining the information by calculation - for example the amount of
water in the tank is obtained from the stick reading and then converted to volume - do this after
completing the test. Indicate the unit of all temperature measurements by checking the
appropriate box.
Topping Off Records, If Applicable
If topping off is not part of the procedure, indicate as NA. Fill in all the blanks.
Conditions At Beginning Of Test
Indicate the date and time when the vendor begins setting up his test equipment. This is not the
start of the test data collection itself.
The evaluator's field crew starts inducing the leak rate and records the time on pages 4 and 5.
Record all leak simulation data using the form on pages 4 and 5.
Once the evaluator's field crew is ready with the induced leak rate simulation and the vendor's
crew starts the actual testing, record the date and time the vendor's test data collection starts.
Also, indicate the product temperature at the time. Fill out the weather condition section of the
form. Indicate the nominal leak rate obtained from the randomization design.
Conditions At Completion Of Testing
Indicate date and time the test is completed.
Again, stick the tank and record the readings and the amount of water in the tank. Record all
weather conditions.
Leak Rate Data
The evaluator's statistician or analyst performing the calculations fills out this section. He can
complete this section as the evaluation proceeds or at the end of the evaluation.
The nominal leak rate is obtained from page 2; see test conditions at beginning of test. Check it
against the nominal leak rate in the randomization design by matching test numbers.
The induced leak rate is obtained by calculation from the data reported by the evaluating field
crew on pages 4 and 5, if needed, of this form. The vendor's crew reports the measured leak
rate.
Calculate the difference by subtracting the induced from the measured leak rate.
B-9

-------
Additional Comments, If Needed
Use this page for comments, such as adverse weather conditions, method failure, and reason for
invalid test pertaining to test.
Induced Leak Rate Data
The evaluator's field crew completes this form on pages 4 and 5. From the randomization
design, the crew will know the targeted nominal leak rate. They will know the induced leak rate
at the end of the test. However, the test procedures require the induced leak rate be within 10
percent of the nominal leak rate.
B-10

-------
Individual Test Log
Volumetric Tank Tightness Testing Method
Name of field operator
Signature of field operator
Test no.	Date
Instructions: Use one log for each test.
Fill in the blanks and check the boxes, as
appropriate. Keep test log even if test is
inconclusive.
1. General Background Information
Method name and version
Product type
Type of tank
Tank dimensions (nominal)
Diameter:	Inches
Length:	Inches
Volume:	Gallons
Groundwater level
	 Inches above bottom of tank
Recommended stabilization period before test,
per vendor SOP
Hours
Minutes
2. Conditions Before Testing
Date and time at start of condition test tank
Date	Military
time
Stick reading before partial emptying of tank
• Product
Inches	Gallons
• Water
Inches
Gallons
Temperature of product in tank before partial
emptying
~ °F ~ °C
Stick reading after partial emptying of tank -
product
	 Inches 	 Gallons
Amount of product removed from tank
determined by subtraction	Gallons
Stick reading after filling to test level
• Product
Inches
Gallons
• Water
Inches
Gallons
Amount of product added to fill tank determined
by subtraction	gallons
Temperature of product added to fill tank
	 ~ °F ~ °C
Temperature of product in tank immediately
after filling
~ °F ~ °C
Date and time at completion of fill
Date
Military
time
3. Topping Off Records, If
Applicable
Date and time at completion of topping off
Date	Military
time
Approximate amount of product added
	Gallons
If tank overfilled, height of product above tank
Inches
Volumetric TTT Method - Individual Test Log
Page 1 of 4

-------
Individual Test Log
Volumetric Tank Tightness Testing Method
Name of field operator
Signature of field operator
Test no.	Date
4. Conditions At Beginning Of Test
Date and time vendor began setting up test
equipment
Date	Military
time
Complete induced leak rate data sheet; use
attached pages 4 and 5
Date of test data collection
Start time of test data collection
	 Military
Temperature of product at start of test
	 ~°ForD°C
Nominal leak rate
	Gallon per hour
• Weather conditions at beginning of test
Date and time at completion of test data
collection
Date	Military
time
Stick reading at completion of test data
collection
• Product
Inches	Gallons
• Water
Inches	Gallons
• Weather conditions at end of test
Ambient temperature
Barometric pressure
~ °F or ~ °C
~ mmHg
I"! inches Hg
Wind
I~1 None
~	Light
1 1 Moderate
~	Strong
Precipitation
I~1 None
~ Light
1 1 Moderate
I"! Heavy
Sky condition
1 1 Sunny Q Cloudy
1 1 Partly cloudy EH Dark
Date and time test method is disassembled, if
done for this test, and tank is ready for service
Date	Military
time
Ambient temperature
Barometric pressure
~ °F or ~ °C
~ mmHg

1 I inches Hg
Wind
Precipitation
1 I None
1 I None
~ Light
~ Light
1 I Moderate
1 I Moderate
~ Strong
1 I Heavy
Sky condition

1 1 Sunny
1 1 Cloudy
1 1 Partly cloudy
~ Dark
Nominal leak rate	gal/hr
5. Conditions At Completion Of
Testing
Volumetric TTT Method - Individual Test Log	Page 2 of 4

-------
Individual Test Log
Volumetric Tank Tightness Testing Method
Name of field operator
Signature of field operator
Test no.	Date
6. Leak Rate Data; Not To Be Filled
Out By Field Crew
Nominal leak rate 	 gal/hr
Induced leak rate 	 gal/hr
Leak rate measured by vendor's method
	 gal/hr
Difference measured rate minus induced rate
	 gal/hr
Additional explanations or comments
Volumetric TTT Method - Individual Test Log
Page 3 of 4

-------
Name of field operator
Signature of field operator	Test no.
Date of test	Induced Leak Rate Data Sheet

Time At
Product
Collection
(military)
Amount Of
Product
Collected
(mL)
Comments, If Applicable
1



2



3



4



5



6



7



8



9



10



11



12



13



14



15



16



17



18



19



20



21



22



23



24



Volumetric TTT Method - Individual Test Log
Page 4 of 4

-------
Appendix C
Non-Volumetric Methods Reporting Forms
C-l

-------
Non-Volumetric Methods Evaluation Forms
Appendix C provides five sets of blank forms. When completed, these forms provide the
framework for a standard report. They consist of:
1.	Results Of U.S. EPA Standard Evaluation - Non-volumetric Tank Tightness Testing
Method
2.	Description - Non-volumetric Tank Tightness Testing Method
3.	Reporting Form For Leak Test Results - Non-volumetric Tank Tightness Testing
Method
4.	Individual Test Log - Non-volumetric Tank Tightness Testing Method
5.	Reporting Form For Water Sensor Evaluation Data - Non-volumetric Tank Tightness
Testing Method
Each set of forms includes instructions on how to fill out the forms and who should complete
them. Below is an overview of various responsibilities.
1.	Results of U.S. EPA Standard Evaluation - The evaluator completes this form at the
end of the evaluation.
2.	Description of Non-volumetric Tank Tightness Testing Method - The evaluator,
assisted by the vendor, completes this form by the end of the evaluation.
3.	Reporting Form For Leak Test Results - The evaluator or the statistician analyzing
the data completes this form. You can develop a blank form on a personal computer,
generate the database for a given evaluation, and merge the two on the computer.
The evaluator can also fill out this form manually. The evaluator's field crew inputs
field test results and vendor's test results on the individual test logs as discussed
below.
4.	Individual Test Logs - The evaluator's field crew completes these forms. Keep these
forms blind to the vendor during testing. The evaluator should reproduce at least 42
copies of the blank form provided in this appendix and produce a bound notebook for
the complete test period.
Non-volumetric methods may require some modification of the test log. We designed
the form in this appendix from a volumetric test log. It is the responsibility of the
evaluator to design the appropriate forms with input from the vendor. It is important
to include in the test logs all parameters relevant to the evaluation of a specific
method. In particular, it is necessary to document the induced leaks.
After completing the evaluation, the evaluator collates all the forms into a single standard report
in the order listed above.
Distribution Of The Evaluation Test Results
The organization performing the evaluation prepares a report to the vendor describing the results
of the evaluation. This report consists primarily of the forms in this appendix. The first form
reports the results of the evaluation. This two-page form is designed to be distributed widely.
C-2

-------
Provide a copy of this form to each tank owner or operator who uses this method of release
detection. The owner or operator must retain a copy of this form as part of his record keeping
requirements. The owner or operator must also retain copies of each tank test performed at his
facility to document the tanks passed the tightness test. Distribute this two-page form to
regulators who must approve release detection methods for use in their jurisdiction.
The evaluator submits the completed report consisting of all the forms in Appendix C to the
vendor of the release detection method. The vendor may distribute the complete report to
regulators who wish to see the data collected during the evaluation. The vendor may also
distribute the report to customers of the release detection method who want to see additional
information before deciding to select a particular release detection method.
The evaluator reports the optional part of the calculations, if conducted, to the vendor of the
release detection method. The vendor may use these calculations to understand the details of the
performance and perhaps improve the method. The vendor can decide whether to distribute this
form.
The evaluator of the release detection method provides the report to the vendor. Distribution of
the results to tank owners or operators and to regulators is the responsibility of the vendor.
C-3

-------
Results Of U.S. EPA Standard Evaluation
Non-volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator fills out this from after completing the evaluation of the method. This form
contains the most important information relative to the method evaluation. Complete all items
and check the appropriate boxes. If a question is not applicable to the method, write NA in the
appropriate space.
This form consists of six main parts:
1.
Method description
2.
Evaluation results
3.
Test conditions during evaluation
4.
Limitations on the results
5.
Certification of results
6.
Additional evaluation results, if applicable
Method Description
Indicate the commercial name of the method, the version, and the name, address, and telephone
number of the vendor. Some vendors might use different versions of their method when using it
with different products or tank sizes. If so, indicate the version used in the evaluation. If the
vendor is not the party responsible for the development and use of the method, then indicate the
home office name and address of the responsible party.
Evaluation Results
Report the evaluation results separately for each detection mode if the method operates in
different detection modes depending on field conditions. Describe the mode of detection for
which the results are applicable.
Calculate P(fa), which is the probability of false alarm.
Report the number of false alarms and the number of tight tank tests, and report the 95 percent
confidence interval based on the binomial distribution with Ni tests.
In the blank, insert the leak rate used in the evaluation. This is the leak rate corresponding to the
reported P(d) below.
Calculate P(d), which is the probability of detecting a leak of the size induced of no more than
0.10 gal/hr.
Report the number of correct detections and the number of simulated leak tests, and report the 95
percent confidence interval based on the binomial distribution with N2 tests.
C-4

-------
If the calculated P(fa) is 5 percent or less and if the calculated P(d) is 95 percent or more, check
the does box. Otherwise, check the does not box. Note: the P(fa) and P(d) requirements apply
to each release detection mode used by the method.
Indicate whether this method operates under more than one mode of detection. Check the
appropriate box and complete page 4 regarding additional evaluation results, if applicable.
Test Conditions During Evaluation
Insert the information in the blanks provided. The nominal volume of the tank in gallons is
requested, as is the tank material of steel or fiberglass. Also, report the backfill material in the
tank excavation, for example clean sand or pea gravel. Give the tank diameter and length in
inches. Report the product used in the testing. Give the range of temperature differences
actually measured, as well as the standard deviation of the observed temperature differences.
Report the groundwater level for the test tank in inches above the bottom of the tank. Report 0
for groundwater at or below the bottom of the tank.
Other sources of interference may affect non-volumetric methods. Report any sources of
interference specific to the method on the lines provided. Also, report the range of test
conditions for the indicated interference source. If no additional sources of interference are
identified, check none.
Limitations On The Results
Where applicable, the size in gallons of the largest tank to which these results can be applied
may be calculated as 1.5 times the size in gallons of the test tank. There are methods, such as
vacuum decay methods, where this is not applicable.
Determine the temperature differential, the waiting time after adding product until testing, and
the total data collection time using the results from calculations. Alternately, if the principle of
operation of the method is unaffected by product temperature changes, check the box indicating
temperature is not a limiting factor and give the justification.
Certification Of Results
The evaluator indicates which test procedure was followed and provides his or her name and
signature, and the name, address, and telephone number of the organization.
Additional Evaluation Results, If Applicable
If checking the yes box relating to other release detection modes on page 1, then provide the
necessary information for the P(fa) and P(d) for the additional release detection mode. Calculate
these probabilities, based on the evaluation results obtained in detection mode.
C-5

-------
Results Of U.S. EPA Standard Evaluation
Non-volumetric Tank Tightness Testing Method
This form tells whether the tank tightness testing method described below complies with the
performance requirements of the federal UST regulation. The vendor or a consultant to the
vendor conducted the evaluation according to U.S. EPA's Standard Test Procedures For
Evaluating Release Detection Methods: Volumetric and Non-volumetric Tank Tightness Testing.
The full evaluation report also includes a form describing the method and a form summarizing
the test data.
Tank owners using this release detection method should keep this form on file to prove
compliance with the federal UST regulation. Tank owners should check with regulatory
authorities to make sure this form satisfies their requirements.
Method Description
Name	
Version	
Vendor	
Street address
City State
Zip
Evaluation Results

This method, which declares a tank to be leaking when

has an estimated probability of false alarms or P(fa) of	% based on the test results of	false alarms out
of	tests. A 95% confidence interval for P(fa) is from
	to	%.
The corresponding probability of detection or P(d) of a _gal/hr leak is	% based on the test results of	
detections out of	simulated leak tests. A 95% confidence interval for P(d) is from
	to	%.
Does this method use additional modes of release detection? ~ yes ~ no. If yes, complete additional
evaluation results on page 3 of this form.
Based on the results above and on page 3 if applicable, this method O does O does not meet the federal
performance standards established by the U.S. Environmental Protection Agency of 0.10 gal/hr at P(d) of
95% and P(fa) of 5%.
Non-volumetric TTT Method - Results Form
Page 1 of 4

-------
Test Conditions During Evaluation
The evaluation testing was conducted in a	- gallon ~ steel ~ fiberglass tank, which was
	inches in diameter and	inches long, installed in	backfill.
The groundwater level was	inches above the bottom of the tank.
Non-volumetric TTT method	
Version	
Test Conditions During Evaluation (continued)
The tests were conducted with the tank	% full.
The temperature difference between product added to fill the tank and product already in the tank ranged
from	°F to	°F, with a standard deviation of	°F.
The product used in the evaluation was	.
This method may be affected by other sources of interference. List these interferences below and give the
ranges of conditions under which the evaluation was done. Check none if not applicable.
I~~l none
Interferences	Range Of Test
Conditions
Limitations On The Results
The performance estimates above are only valid when
•	The method has not been substantially changed.
•	The vendor's instructions for using the method are followed.
•	The tank contains a product identified on the method description form.
•	The tank capacity is	gallons or smaller.
•	The difference between added and in-tank product temperatures is no greater than + or - _degrees
Fahrenheit.
•	This method can be used on up to	of tanks connected by siphon piping with a total volume
of	.
~ check if applicable
Temperature is not a factor because	
Non-volumetric TTT Method - Results Form
Page 2 of 4

-------
•	The waiting time between the end of filling the test tank and the start of the test data collection is
at least	hours.
•	The waiting time between the end of topping off to final testing level and the start of the test data
collection is at least	hours.
•	The total data collection time for the test is at least	hours.
•	The product volume in the tank during testing is	% full.
•	This method ~ can ~ cannot be used if the groundwater level is above the bottom of the tank.
Other limitations specified by the vendor or determined during testing
Non-volumetric TTT method	
Version	
Safety disclaimer: This test procedure only addresses the issue of the method's ability to detect leaks. It
does not test the method for safety hazards.
Additional Evaluation Results, If Applicable
This method, which declares a tank to be leaking when	
has an estimated probability of false alarms or P(fa) of	% based on the test results of	false
alarms out of	tests. Note: A perfect score during testing does not mean the method is perfect.
Based on the observed results, a 95% confidence interval for P(fa) is from 0 to	%.
The corresponding probability of detection or P(d) of a	gal/hr leak is	% based on the test results
of	detections out of	simulated leak tests. Note: A perfect score during testing does not mean
the method is perfect. Based on the observed results, a 95% confidence interval for P(d) is from 0 to
0/
	/O.
Water Detection Mode, If Applicable
Using a false alarm rate of 5%, the minimum water level the water sensor can detect with a 95%
probability of detection is	inches.
Using a false alarm rate of 5%, the minimum change in water level the water sensor can detect with a
95% probability of detection is	inches.
Based on the minimum water level and change in water level the water sensor can detect with a false
alarm rate of 5% and a 95% probability of detection, the minimum time for the method to detect an
increase in water level at an incursion rate of 0.10 gal/hr is	minutes in a	gallon
tank.
Non-volumetric TTT Method - Results Form
Page 3 of 4

-------
Certification Of Results
I certify the non-volumetric tank tightness testing method was installed and operated according to the
vendor's instructions. I also certify the evaluation was performed according to the standard EPA test
procedure for non-volumetric tank tightness testing methods and the results presented above are those
obtained during the evaluation.
Printed name	Organization performing evaluation
Signature	City, state, zip
Date	Phone number
Non-volumetric TTT Method - Results Form
Page 4 of 4

-------
Description Of Non-volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator, with assistance from the vendor, fills out this form, as part of the evaluation of the
method. This form provides supporting information on the principles behind the method or on
how the method works.
To minimize the time to complete this form, we provide possible answers to the most frequently
expected questions. For those answers dependent on site conditions, give answers that apply in
typical conditions. Write in any additional information about the testing method you believe is
important.
There are seven parts to this form. They are:
1.	Method Name and Version
2.	Product
•	Product type
•	Product level
3.	Principle of Operation
4.	Temperature Measurement
5.	Data Acquisition
6.	Procedure Information
•	Waiting times
•	Test duration
•	Total time
•	Other important elements of the procedure or method
•	Identifying and correcting for interfering factors
•	Interpreting test results
7.	Exceptions
Indicate the commercial name and the version of the method in the first part.
Note: The version is provided for methods that use different versions of the method for different
products or tank sizes.
For the six remaining parts, check all appropriate boxes for each question. Check more than one
box per question if it applies. If a box other is checked, complete the space provided to specify
or briefly describe the matter. If necessary, use all the white space next to a question for a
description.
Complete the section about other important elements of the procedure or method completed
carefully. List here any other important elements of the method that could affect its
performance. For example:
C-6

-------
•	If the pressure in the ullage space is different from atmospheric during testing, indicate
whether a negative or positive pressure was applied. Report pressure and its units.
•	If the method used is a tracer method, clearly document the process of adding the tracer
to the tank and in the spiking port.
•	If a tracer is added to the product in the tank, provide information on these items:
o type of tracers
o tracer concentration in the product
o type of carrier
o time between spiking and starting the test
o type of sampling, for example, whether sampling is active or passive; in other
words, how does the tracer reach the sampling ports? by natural diffusion
process? is the process enhanced by adding forced air?
o other relevant items
•	When sampling ports are installed for tracer methods, measure the distances between any
parts of the tank to its nearest sampling port. Report the largest of these distances.
C-7

-------
Description
Non-volumetric Tank Tightness Testing Method
This section describes briefly the important aspects of the non-volumetric tank tightness testing
method. It is not intended to provide a thorough description of the principles behind the method
or how the method works.
Method Name And Version
Product
Product type
For what products can this method be used? Check all that apply.
I~~l gasoline
l~~l diesel
I I aviation fuel
l~~l fuel oil #4
l~~l fuel oil #6
I I solvents
~	waste oil
~	other, list	
Product level
What product level is required to conduct a test?
I I	above grade
l~~l	within the fill pipe
l~~l	greater than 90% full
l~~l	greater than 50% full
~	empty
~	other, specify	
Principle Of Operation
What principle or principles are used to identify a leak?
~	acoustical signal characteristic of a leak
~	identification of a tracer chemical outside the tank system
l~~l changes in product level or volume
~	detection of water inflow
~	other, describe briefly	
Non-volumetric TTT Method - Description
Page 1 of 5

-------
Temperature Measurement
If product temperature is measured during a test, how many temperature sensors are used?
I I single sensor, without circulation
l~~l single sensor, with circulation
l~~l 2-4 sensors
Q 5 or more sensors
~	temperature-averaging probe
If product temperature is measured during a test, what type of temperature sensor is used?
I I resistance temperature detector (RTD)
l~~l bimetallic strip
~	quartz crystal
~	thermistor
~	other, describe briefly	
If product temperature is not measured during a test, why not?
~	the factor measured for change in level or volume is independent of temperature, for example
mass
I I the factor measured for change in level or volume self-compensates for changes in
temperature
~	other, explain briefly	
Data Acquisition
How are the test data acquired and recorded?
I~~l manually
l~~l by strip chart
l~~l by computer
Procedure Information
Waiting times
What is the minimum waiting period between adding a large volume of product to bring the level to test
requirements and the beginning of the test, for example from 50% to 95% capacity?
I I not applicable
l~~l no waiting period
~	less than 3 hours
~	3-6 hours
l~~l 7-12 hours
~	more than 12 hours
~	variable, depending on tank size, amount added, and operator discretion
Non-volumetric TTT Method - Description
Page 2 of 5

-------
Test duration
What is the minimum time for collecting data?
~	less than 1 hour
l~~l 1 hour
l~~l	2 hours
l~~l	3 hours
l~~l	4 hours
l~~l	5-10 hours
~	more than 10 hours
~	variable
Total time
What is the total time needed to test with this method?
Calculate setup time plus waiting time plus testing time plus time to return tank to service.
	hours	minutes
Other important elements of the procedure or method
List other elements that could affect the performance of the procedure or method; for example, positive or
negative ullage pressure, tracer concentration, and distance between tank and sampling ports
Identifying and correcting for interfering factors
How does the method determine the presence and level of the groundwater above the bottom of the tank?
~	observation well near tank
I I information from USGS or others
I I information from personnel on-site
I I presence of water in the tank
l~~l other, describe briefly	
~	level of groundwater above bottom of the tank not determined
How does the method correct for the interference due to the presence of groundwater above the bottom of
the tank?
I~~l head pressure increased by raising the level of the product
I I different head pressures tested and leak rates compared
l~~l tests for changes in water level in tank
Non-volumetric TTT Method - Description
Page 3 of 5

-------
~	other, describe briefly	
~	no action
Does the method measure inflow of water as well as loss of product in gal/hr?
~	yes
~	no
Does the method detect the presence of water in the bottom of the tank?
~	yes
~	no
How does the method identify the presence of vapor pockets?
~	erratic temperature, level, or temperature-compensated volume readings
~	sudden large changes in readings
~	statistical analysis of variability of readings
l~~l other; describe briefly	
I I not identified
I I not applicable, under filled test method used
How does the method correct for the presence of vapor pockets?
I I bleed off vapor and start test over
I I identify periods of pocket movement and discount data from analysis
l~~l other, describe briefly	
l~~l not corrected
~	not applicable, under filled test method used
Are the method's sensors calibrated before each test?
~	yes
~	no
If not, how often are the sensors calibrated?
I I weekly
I I monthly
I I yearly or less frequently
l~~l never
Interpreting test results
What effect is used to declare the tank to be leaking? List all modes used by the method.
Non-volumetric TTT Method - Description
Page 4 of 5

-------
If a change in volume is used to detect leaks, what threshold value for product volume change in gal/hr is
used to declare a tank is leaking?
I~~l 0.05 gal/hr
l~~l 0.10 gal/hr
l~~l 0.20 gal/hr
I I other	
Under what conditions are test results considered inconclusive?
I~~l groundwater level above bottom of tank
~	presence of vapor pockets
I I too much variability in the data with standard deviation beyond a given value
l~~l unexplained product volume increase
l~~l other, describe briefly	
Exceptions
Are there any conditions under which a test should not be conducted?
I~~l groundwater level above bottom of tank
l~~l presence of vapor pockets
I I large difference between ground temperature and delivered product temperature
I I extremely high or low ambient temperature
I I invalid for some products, specify	
~	soil not sufficiently porous
~	other, describe briefly	
What are acceptable deviations from the standard testing test procedure?
I~~l none
l~~l lengthen the duration of test
~	other, describe briefly	
What elements of the test procedure are left to the discretion of the testing personnel on site?
~	waiting period between filling tank and beginning test
I I length of test
~	determination of presence of vapor pockets
~	determination of outlier data may be discarded
l~~l other, describe briefly	
l~~l none
Non-volumetric TTT Method - Description
Page 5 of 5

-------
Reporting Form For Leak Test Results
Non-volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator fills out this form after completing the evaluation of the method in each of its
release detection modes. This form provides for 60 test results, although the minimum number
of tests required in the test procedure is 42. Use as many pages as necessary to summarize all of
the tests attempted. Report the results for each release detection mode on separate forms.
Indicate the commercial name and the version of the method and the period of evaluation above
the table. The version is provided for methods that might use different versions of the method
for different products or tank sizes. Also, indicate the release detection mode for which these
results were obtained.
In general, the statistician analyzing the data completes this form. You may develop a blank
form on a personal computer, generate the database for a given evaluation, and merge the two on
the computer. You can also complete this form manually. The input for the form consists of the
field test results recorded by the evaluator's field crew on the individual test logs and the
vendor's test results.
The table consists of 10 columns. One line is provided for each test performed during evaluation
of the method. If a test is invalid or aborted, list the test with the appropriate notation, for
example invalid on the line.
The test number in the first column refers to the test number from the randomization design
determined according to the test procedures. Since some changes to the design might occur
during the course of the field-testing, the test numbers might not always be in sequential order.
Report the results from the trial run need here as well.
The following list matches the column input required with its source, for each column in the
table.
Column No.	Input	Source
1
2
3
4
5
6
7
Test number or trial run
Date at completion of last fill, if applicable
Time at completion of last fill, if applicable
Date test began
Time test began
Time test ended
Product temperature differential, if
Randomization design
Individual test log
Individual test log
Individual test log
Individual test log
Individual test log
Individual test log
9
10
8
applicable
Nominal leak rate
Induced leak rate
Leak test results
Randomization design
Individual test log
Vendor's test result
C-8

-------
Note: The product temperature differential in column 7 is the difference between the
temperature of the product added and of the product in the tank, each time the tank is filled. This
temperature differential is the actual differential achieved in the field and not the nominal
temperature differential.
C-9

-------
Reporting Form For Leak Test Results
Non-volumetric Tank Tightness Testing Method
Method name and version		Release detection mode
Evaluation period from	to	(dates)

If Applicable
If Applicable

If Applicable


Date At
Completion
Of Last Fill
(m/d/y)
Time At
Completion
Of Last Fill
(military)
Date Test
Began
(m/d/y)
Time Test
Began
(military)
Time Test
Ended
(military)
Product
Temperature
Differential
(° F)
Nominal
Leak Rate
(gal/hr)
Induced
Leak Rate
(gal/hr)
Tank Tight?
(Yes, No, Or
Test Invalid)
Test No.
Trial Run





0
0
0











1









2









3









4









5









6









7









8









9









10









11









12









13









14









15









16









17









18









19









20









Non-volumetric TTT-Data Reporting Form
Page 1 of 3

-------
Reporting Form For Leak Test Results
Non-volumetric Tank Tightness Testing Method
Method name and version		Release detection mode
Evaluation period from	to	(dates)

If Applicable
If Applicable

If Applicable


Date At
Completion
Of Last Fill
(m/d/y)
Time At
Completion
Of Last Fill
(military)
Date Test
Began
(m/d/y)
Time Test
Began
(military)
Time Test
Ended
(military)
Product
Temperature
Differential
(° F)
Nominal
Leak Rate
(gal/hr)
Induced
Leak Rate
(gal/hr)
Tank Tight?
(Yes, No, Or
Test Invalid)
Test No.
21









22









23









24









25









26









27









28









29









30









31









32









33









34









35









36









37









38









39









40









Non-volumetric TTT-Data Reporting Form
Page 2 of 3

-------
Reporting Form For Leak Test Results
Non-volumetric Tank Tightness Testing Method
Method name and version		Release detection mode
Evaluation period from	to	(dates)

If Applicable
If Applicable

If Applicable


Date At
Completion
Of Last Fill
(m/d/y)
Time At
Completion
Of Last Fill
(military)
Date Test
Began
(m/d/y)
Time Test
Began
(military)
Time Test
Ended
(military)
Product
Temperature
Differential
(° F)
Nominal
Leak Rate
(gal/hr)
Induced
Leak Rate
(gal/hr)
Tank Tight?
(Yes, No, Or
Test Invalid)
Test No.
41









42









43









44









45









46









47









48









49









50









51









52









53









54









55









56









57









58









59









60









Non-volumetric TTT-Data Reporting Form
Page 3 of 3

-------
Individual Test Log
Non-volumetric Tank Tightness Testing Method
Instructions For Completing The Form
The evaluator's field crew completes the test log form. Fill out a separate form for each
individual test including the trial run; that means at least 43. Keep the information on these
forms blind to the vendor during the period of evaluation of the method. Adapt the form as
needed to document the evaluation data.
The form consists of nine parts:
1.
Header information
2.
General background information
3.
Conditions before testing
4.
Topping off records, if applicable
5.
For tracer methods only
6.
Conditions at beginning of test
7.
Conditions at completion of testing
8.
Leak rate data
9.
Additional comments, if needed
10.
Data sheet for leak simulation for tracer methods
11.
Data sheet for induced leak rate calibration
All items are to be filled out and the appropriate boxes checked. If a question is not applicable,
then indicate as NA. The following provides guidance on the use of this form.
Header Information
Repeat the header information on all five pages, if used. If a page is not used, cross it out and
initial it. The evaluator's field operator must print and sign his or her name and note the date of
the test on top of each sheet.
Obtain the test number from the randomization design. It is not the sequential running test
number. If a test must be rerun, indicate the test number of the test being rerun and indicate that
on the test log, for example, test no. 5 repeat.
General Background Information
Indicate the commercial name of the method. Include version identification if the method uses
different versions for different products or tank sizes. Prior to testing, obtain the recommended
stabilization period, if applicable, from the vendor. This is important since it influences
scheduling the evaluation. All other items in this section refer to the test tank and product.
Indicate the groundwater level at the time of the test.
C-10

-------
Theoretically, this information remains unchanged for the whole evaluation period. However,
changing weather conditions can affect the groundwater level or the evaluator can change the test
tank.
Conditions Before Testing
Fill in all the blanks. If the information is obtained by calculation, this can be done after the test
is completed. An example is the amount of water in the tank is obtained from the stick reading
and then converted to volume. Indicate the unit of all temperature measurements by checking
the appropriate box.
Note: The term conditioning refers to all activities undertaken by the evaluating field crew to
prepare for a test. As such, the term refers to emptying or filling the tank, heating or cooling
product, and changing the leak rate. In some cases, all of the above are performed; in others,
only one parameter might change. For tracers, conditioning refers to preparation of the tank for
testing. It includes determining the time to wait between spiking and testing.
Topping Off Records, If Applicable
If you perform this step, fill in the appropriate blanks.
For Tracer Methods Only
Fill in the appropriate information. Follow the instructions and complete the form on page 4.
Conditions At Beginning Of Test
The evaluation organization's field crew calibrates the leak simulation equipment prior to the
test. Document all leak rate calibration data need using the form on pages 4 or 5, as appropriate.
Refer to previous calibration if done previously. Adapt the form as necessary.
Once the evaluator's field crew has the induced leak rate simulation, and the vendor starts the
actual testing, record the date and time the vendor's test data collection starts. Also, indicate the
product temperature at the time. Fill out the weather condition section of the form. Indicate the
nominal leak rate, obtained from the randomization design.
Conditions At Completion Of Testing
Indicate date and time when the test is completed.
Again, stick the tank and record the readings and amount of water in the tank. Record all
weather conditions as requested.
C-ll

-------
Leak Rate Data
The evaluator's statistician or analyst performing the calculations fills out this section.
Therefore, he can complete it as the evaluation proceeds or at the end of the evaluation.
Obtain the nominal leak rate from page 2 under conditions at beginning of test. Check the rate
against the nominal leak rate in the randomization design by matching test numbers.
Obtain the induced leak rate from the simulation data reported by the evaluating field crew on
pages 4 or 5 of this form.
The vendor obtains the test result.
Identify the mode on the line following the test answer if the method uses more than one mode of
release detection.
Additional Comments, If Needed
Use this page for any comments pertaining to the test. Examples include adverse weather
conditions, method failure, and reason for invalid test.
Leak Simulation Form For Tracer Methods
For tracer methods, use the form on page 4 to document and measure delivery of the carrier with
the appropriate concentration of the tracer to the spiking ports. Indicate the tracer used and the
concentration of tracer in the carrier in the appropriate spaces. Report the distances between
spiking port and all sampling ports. Record the time and amount of material released in the
spiking port to document the leak simulation for tracer methods. Use as many pages as needed.
Induced Leak Rate Calibration Form
For acoustical methods, use the form on page 5 to calibrate the liquid flow through the simulator
under a standard set of conditions. The induced leak rate is the rate at which the liquid will flow
at a specified head or depth of product. Determine this rate by calibration and use it as the leak
rate for detection. Perform the calibration at a different time than and preferably before, the
testing. Calibrate for each distinct leak rate. After completing the calibrations, document on
each daily test log the simulation conditions and reference the appropriate calibration data sheets,
which should be attached to the daily test log that first uses the given induced leak rate.
C-12

-------
Individual Test Log
Non-volumetric Tank Tightness Testing Method
Name of field operator	
Signature of field operator	 Test no.
Date of test	
Instructions
Use one log for each test.
Fill in the blanks and check the boxes, as appropriate.
Keep test log even if test is inconclusive.
General Background Information
Method name and version	
Product type	
Type of tank	
Tank dimensions nominal measurement
Diameter	inches
Length	inches
Volume	inches
Groundwater level	inches above bottom of tank
Recommended stabilization period before test, per vendor standard operating procedure
	hours	minutes
Conditions Before Testing
Date	and military time	at start of conditioning test tank
Stick reading before partial emptying of tank
Product	inches	gallons
Water	inches	gallons
Temperature of product in tank before partial emptying 	 	°fD or °C ~
Stick reading after partial emptying of tank
Product	inches	gallons
Amount of product removed from tank using subtraction	gallons
Stick reading after filling tank to test level
Product	inches 	gallons
Water	inches 	gallons
Amount of product added to fill tank using subtraction	gallons
Non-volumetric TTT - Data Log
Page 1 of 5

-------
Name of field operator	
Signature of field operator		Test no.
Conditions Before Testing (Continued)
Temperature of product added to fill tank	 	°F ~ or °C ~
Temperature of product in tank immediately after filling	 	°fD or °C ~
Date	and military time	at completion of fill
Topping Off Records, If Applicable
Date	and military time	
Approximate amount of product added	
If tank overfilled, height of product above tank
For Tracer Methods Only
Date	and military time	tracers added to product in test tank
Tracer used	
Amount of tracer used	
Amount of product in test tank	gallons
Complete The Tracer Leak Simulation Form, Use Page 4
Date	and military time	at start of test
Date	and military time	at conclusion of test
Conditions At Beginning Of Test
Date	and military time	vendor began setting up test equipment
Document Induced Leak Rate Determination, Use Page 5
Date	and military time	at start of vendor's test data collection
Temperature of product in tank at start of test 	 	°fD or °C ~
Weather conditions
Temperature	 	°fD or °C ~
Barometric pressure	mm Hg ~ or	in. Hg ~
Wind	None ~	Light ~	Moderate ~
Precipitation	None ~	Light ~	Moderate ~
Sunny ~	Partly cloudy ~	Cloudy ~
Nominal leak rate	gal/hr
at completion of topping off
.gallons
	inches
Strong ~
Heavy I I
Non-volumetric TTT - Data Log
Page 2 of 5

-------
Name of field operator	
Signature of field operator		Test no.
Conditions At Completion Of Testing
Date	and military time	at completion of test data collection
Stick reading at completion of test data collection
Product	inches	gallons
Water	inches	gallons
Date of test	
Conditions At Completion Of Testing (Continued)
Temperature of product in tank at start of test 	 	°F ~ or °C ~
Weather conditions
Temperature	 	°F ~ or °C ~
Barometric pressure	mm Hg ~ or	in. Hg ~
Wind	None ~	Light ~	Moderate Q	Strong ~
Precipitation	None ~	Light ~	Moderate ~	Heavy ~
Sunny ~	Partly cloudy ~	Cloudy ~
Date	and military time	test method is disassembled, if done for this test, and tank is ready for
service
Leak Rate Data
Release detection mode	
Nominal leak rate	gal/hr
Induced leak rate	gal/hr
Findings for tracer methods
~ No tracer found ~ Tracers found
If tracers found, list 	
Test answer Q Leaking ~ Tight ~ Inconclusive
Additional Comments, Use Back Of Page If Needed
Non-volumetric TTT - Data Log	Page 3 of 5

-------
Name of field operator	
Signature of field operator
Date of test		Test no.
Leak Simulation Form For Tracer Method
Reproduce Form, If Needed
Tracer used.
Carrier	
Concentration of tracer in carrier	
Distance from spiking port to
Sampling port 1		Sampling port 5_
Sampling port 2		Sampling port 6_
Sampling port 3		Sampling port 7_
Sampling port 4		Sampling port 8_
Non-volumetric TTT - Data Log
Page 4 of 5

-------
Name of field operator
Signature of field operator	
Date of test		Test no.
Induced Leak Rate Calibration Form
Reproduce Form, If Needed

Time
(military)
Amount*
Comments
1



2



3



4



5



6



7



8



9



10



11



12



13



14



15



16



17



18



19



20



21



22



23



24



* Indicate all measurement units.
Non-volumetric TTT - Data Log
Page 5 of 5

-------
Appendix D
Sensor Evaluation Forms
D-l

-------
Results Of U.S. EPA Alternative Evaluation
Sensors
This form documents the performance of the sensor described below. The vendor, or a consultant to the vendor,
conducts the evaluation according to the U.S. EPA's requirements for alternative protocols. The full evaluation
report includes a report describing the method, a description of the evaluation procedures, and a summary of the
test data.
Tank owners using this release detection system should keep this form on file to prove compliance with the
federal UST regulation. Tank owners should check with regulatory authorities to make sure this form satisfies
their requirements.
Method Description
Name	
V ersion	
Vendor	
Street address
City	State	Zip
Sensor output type	
Sensor operating principle	
General description of the sensor	
Evaluation Results
The sensor listed above was tested for its ability to respond to a change in condition when tested in a controlled
test vessel. The following parameters were determined from this evaluation.
•	Precision standard deviation - Agreement between multiple measurements of the same product level.
•	Detection time - Amount of time the detector must be exposed to product before it responds.
•	Recovery time - Amount of time before the detector stops responding after being removed from the
product.
•	Specificity - Types of products that the sensor will respond to.
Parameter
Ethanol-blended Gasoline
( %)
Water
Diesel
Average detection height in inches



Precision in inches



Average detection time as hh:mm:ss



Recovery time as hh:mm:ss



Sensor Results Form
Page 1 of2

-------




Specificity



Limitations On The Results
Limitations specified by the vendor or determined during testing
Certification Of Results
I certify the sensor was operated according to the vendor's instructions. I also certify the evaluation was performed
according to the standard EPA test procedure for tank tightness testing methods and the results presented above are those
obtained during the evaluation.
Printed name	Organization performing evaluation
Signature	City, state, zip
Date	Phone number
Sensor Results Form
Page 2 of 2

-------
Reporting Form For Liquid Sensor Evaluation Data
Method name and version	
Date of test		Name of field operator
Product type		Signature of field operator

Volume Of
Calculated Liquid

Increment Difference

Liquid Added
Height Increment, h
Sensor Reading
Calc-Meas.
Test No.
(mL)
(in)
(in)
(in)
A
B
C
D
C-E
Minimum Level Detected, X:

inches

1




2




3




4




5




6




7




8




9




10




Note: This form provides a template for data reporting. Use as many pages as necessary.
Sensor Data Log
Page 1 of2

-------
Reporting Form For Vapor Sensor Evaluation Data
Method name and version
Date of test	
Vapor type	
Name of field operator	
Signature of field operator

Baseline Test
Detection Time/
Recovery Time
250 PPM
Detection Time/
Recovery Time
500 PPM
Detection Time/
Recovery Time
1000 PPM
Detection Time/
Recovery Time
Test No.
A
Minimum Level Detected, X:
1




2




3




4




5




6




7




8




9




10




Note: This form provides a template for data reporting. Use as many pages as necessary, one per vapor type.
Sensor Data Log
Page 2 of 2

-------
United States	Land And	EPA 510-B-19-003
Environmental	Emergency Management	May 2019
Protection Agency	5401R	www.epa.gov/ust

-------