US Environmental Protection Agency Office of Pesticide Programs Office of Pesticide Programs Microbiology Laboratory Environmental Science Center, Ft. Meade, MD Standard Operating Procedure for Calibration and Maintenance of Timers SOP Number: EQ-05-08 Date Revised: 08-07-19 ------- SOP No. EQ-05-08 Date Revised 08-07-19 Page i of 4 SOP Number EQ-05-08 Title Calibration and Maintenance of Timers Revisions Made • Added the option of replacing timers when their ISO accreditation expires with newly purchased and accredited timers. • Minor editorial changes. ------- SOP No. EQ-05-08 Date Revised 08-07-19 Page 1 of 4 SOP Number EQ-05-08 Title Calibration and Maintenance of Timers Scope Describes the process for calibration and maintenance of timers, certified instruments that can display time in seconds. Application Timers are used for managing timed activities in test methods. Approval Date SOP Developer: Print Name: SOP Reviewer Print Name: Quality Assurance Unit Print Name: Branch Chief Print Name: Date SOP issued: Controlled copy number: Date SOP withdrawn: ------- SOP No. EQ-05-08 Date Revised 08-07-19 Page 2 of 4 TABLE OF CONTENTS Contents Page Number 1. DEFINITIONS 3 2. HEALTH AND SAFETY 3 3. PERSONNEL QUALIFICATIONS AND TRAINING 3 4. INSTRUMENT CALIBRATION 3 5. SAMPLE HANDLING AND STORAGE 3 6. QUALITY CONTROL 3 7. INTERFERENCES 3 8. NON-CONFORMING DATA 3 9. DATA MANAGEMENT 3 10. CAUTIONS 3 11. SPECIAL APPARATUS AND MATERIALS 3 12. PROCEDURE AND ANALYSIS 3 13. DATA ANALYSIS/CALCULATIONS 4 14. FORMS AND DATA SHEETS 4 15. REFERENCES 4 ------- SOP No. EQ-05-08 Date Revised 08-07-19 Page 3 of 4 1. Definitions 1. A2LA = American Association for Laboratory Accreditation 2. ISO = International Organization for Standardization 3. NVLAP = National Voluntary Laboratory Accreditation Program 2. Health and Safety None 3. Personnel Qualifications and Training Refer to SOP ADM-04, OPP Microbiology Laboratory Training. 4. Instrument Calibration Not applicable 5. Sample Handling and Storage Not applicable 6. Quality Control For quality control purposes, the required information is documented on the appropriate form(s) (see section 14). 7. Interferences None 8. Non- conforming Data 1. Procedures to handle non-conformances are consistent with SOP ADM- 07, Non-Conformance Reports. 9. Data Management 1. Data will be archived consistent with SOP ADM-03, Records and Archives. 10. Cautions Not applicable 11. Special Apparatus and Materials Timer Specifications: • Battery operated bench top timers • Reads in hours, minutes, and seconds • Individual serial numbers • Calibrated per ISO 17025 12. Procedure and Analysis Calibration certificates must contain the stamp of the accrediting body (e.g., A2LA, NVLAP) and the calibration vendor certificate number. 12.1 Use Period of Timers a. Purchase new timers with accredited calibration certificates. b. Maintain an electronic inventory of timers with their calibration certification (Timer Certification Record; see section 14). Enter the timer serial number and calibration expiration date into the electronic inventory. After each addition, file a hard copy of the inventory and calibration certificate in the Timer Calibration Record notebook. ------- SOP No. EQ-05-08 Date Revised 08-07-19 Page 4 of 4 c. When the calibration expires, remove the timer from service and the inventory. 12.2 Operation Follow the instructions, provided by the manufacturer, for the operation of each timer. See section 15. 13. Data Analysis/ Calculations None 14. Forms and Data Sheets Form is stored separately from the SOP under the following file name: Timer Certification Record EQ-05-08 Fl.docx 15. References Operation manuals for timers located in file cabinet in D-wing. ------- |