ORGANOCHLORINE PESTICIDES AND POLY CHLORINATED BIPHENYLS (PCBs) SW-846 Method 8081 or 8080 Table 1A. Summary of Holding' Times and. Preservation for Organochlorine Pesticides and Poly chlorinated Biphenyls (PCBs) Analytical Parameter a Technical and Contract Holding Times Preservation Organochlorine Pesticides and Polychlorinated Biphenyls (PCBs) in Water Samples Technical for Extraction: Cool to 4 ° C ±2 ° C 7 days from collection; Contract for Extraction: 5 days from receipt at laboratory Technical and Contract for Analvsis: 40 davs from extraction Organochlorine Pesticides and Polychlorinated Biphenyls (PCBs) in Soil Samples Technical for Extraction: Cool to 4 ° C ±2 ° C 14 days from collection; Contract for Extraction: 10 days from receipt at laboratory Technical and Contract for Analvsis: 40 davs from extraction a Individual target compounds are listed in Table IB. Data Calculations and Reporting Units: Calculate the calibration factors (CF) of single component pesticides according to Section 7.4.2 of SW-846 Method 8000A. Calculate sample results using the analyte CFs from the midpoint standard of the associated initial calibration curve. Perform sample quantitation for multiple components pesticides according to Section 7.6 of SW-846 Method 8080A or 8081. Report water sample results in concentration units of micrograms per liter (/ig/L) . Report soil sample results on a dry-weight basis in micrograms per kilogram (/ig/kg) . For rounding results, adhere to the following rules: a)If the number following those to be retained is less than 5, round down; b)If the number following those to be retained is greater than 5, round up; or c)If the number following the last digit to be retained is equal to 5, round down if the digit is even, or round up if the digit is odd. All records of analysis and calculations must be legible and sufficient to recalculate all sample concentrations and QC results. Include an example calculation in the data package. 8081CRF 1 of 5 Revision 12/03/1999 ------- TABLE IB. Target Compound. List, CAS Numbers, and Contract Required Quantitation Limits (CRQL) for SW-846 Method 8081 or Method 8080 COMPOUND CAS No. CRQL Water pg/L CRQL Soil pg/kg alpha-BHC 319-84-6 0 . 05 2 beta-BHC 319-85-7 0 . 05 2 delta-BHC 319-86-8 0 . 05 2 gairana-BHC (Lindane) 58-89-9 0 . 05 2 Heptachlor 76-44-8 0 . 05 2 Aldrin 309-00-2 0 . 05 2 Heptachlor epoxide 1024-57-3 0 . 05 2 Endosulfan I 959-98-8 0 . 05 2 Dieldrin 60-57-1 0 .1 3 4,4'-DDE 72-55-9 0 .1 3 Endrin 72-20-8 0 .1 3 Endosulfan II 33213-65-9 0 .1 3 4,4'-DDD 72-54-8 0 .1 3 Endosulfan sulfate 1031-07-8 0 .1 3 4,4'-DDT 50-29-3 0 .1 3 Methoxychlor 72-43-5 0 . 5 17 Endrin ketone 53494-70-5 0 .1 3 Endrin aldehyde 7421-93-4 0 .1 3 alpha-Chiordane 5103-71-9 0 . 05 2 gamma-Chiordane 5103-74-2 0 . 05 2 Toxaphene 8001-35-2 5 170 Aroclor-1016 12674-11-2 1 33 Aroclor-1221 11104-28-2 2 67 Aroclor-1232 11141-16-5 1 33 Aroclor-1242 53469-21-9 1 33 Aroclor-12 48 12672-29-6 1 33 Aroclor-12 54 11097-69-1 1 33 Aroclor-12 60 11096-82-5 1 3 8081CRF 2 of 5 Revision 12/03/1999 ------- 8081CRF 3 of 5 Revision 12/03/1999 ------- Table 2. Summary of Calibration Procedures for Organochlorine Pesticides and Polychlorinated Biphenyls (PCBs) by SW-846 Method 8081 or 8080 Calibration Element Frequency Acceptance Criteria Corrective Action Initial Calibration (minimum blank + 3 points for each analyte) (ICAL) a' b,c Initially; whenever required, due to failure of CCV RSD for CFs <20% (<30% for Surrogate compounds) 1. Terminate analysis 2. Re-calibrate and verify before sample analysis Continuing Calibration Verification (CCV) at midpoint of ICAL Beginning of each day, after every 10 samples, and end of run %D between CF of CCV and avg CFs from ICAL <2 5% 1. Re-calibrate and verify 2. Re-analyze samples back to last good CCV Endrin and 4,4'-DDT Breakdown Beginning and end of analytical sequence <20% each or <30% combined 1. Investigate source of the problem and document 2. If either Endrin, 4,4'-DDT, or their breakdown products were detected, re- analyze the samples a The ICAL low standard must be above but near the CRQL. The low ICAL standard must have a signal to noise ratio >5:1. If this requirement cannot be met, the laboratory must submit a MDL study as part of the data package. b ICAL Prepare initial calibration individual standard mixtures A and B (IND A and IND B) containing the single component pesticides specified in Table 9 of SW-846 Method 8081 at three concentration levels. For multiple response pesticides, including toxaphene and Aroclors (except 1016 and 1260), prepare separate initial calibration standards at the following concentration levels: Aroclors (except 1221)at 100 ng/mL; Aroclor-1221 at 200 ng/mL; and toxaphene at 500 ng/mL. Aroclor-1016 and Aroclor-1260 may be combined into a single standard solution. Spike all calibration standards with the surrogate compounds discussed in Table 3 at a concentration of 20 ng/mL. c Report the retention time window for each analyte. For multiple component pesticides, calculate the retention time window for 5 major peaks from the initial calibration standard analysis. Determine retention time windows for both single and multiple component pesticides using the following guidelines: 8081CRF 4 of 5 Revision 12/03/1999 ------- Column Type Packed Column Mega bore or wide bore capillary column Retention Time Window in Minutes <± 2=8 ±0.05 for tetrachloro-m-xylene through Aldrin ±0.07 for compounds which elute after Aldrin ±0.1 for decachlorobiphenyl 8081CRF 5 of 5 Revision 12/03/1999 ------- Table 3. Summary of Internal Quality Control Procedures for Organochlorine Pesticides and Polychlorinated Biphenyls (PCBs) by SW-846 Method 8081 or 8080 QC Element Frequency Acceptance Criteria Corrective Action Method Blank (MB) One per Batch or SDGa (1 per 20 samples minimum) < CRQL for each compound 1. Investigate source of contamination and document 2. Re-extract and re-analyze all samples processed with a non-compliant method blank Surrogate b Every standard, sample, method blank and QC sample at 10 times CRQL 60-150% of expected value 1. Re-analyze all samples with non-compliant surrogate recoveries Matrix Spike and Matrix Spike Duplicate (MS/MSD)c One MS/MSD set per batch or SDG (1 MS/MSD set per 20 samples minimum) 50-135% of expected value; <30 RPD between MS and MSD 1. Address in narrative a SDG - Sample Delivery Group - each case of field samples received; or each 20 field samples within a case; or each 14 calendar day period during which field samples in a case are received. b Spike each standard, sample, and blank with lmL of a solution containing 0.2 /ig/mL each of tetrachloro-m- xylene and decachlorobiphenyl c Spike MS/MSD samples with lmL of a solution containing the following compounds and levels: Target compound Concentration tua/mLl Target Compound Concentration tua/mLl ?-BHC 0.5 Heptachlor 0.5 4, 4'-DDT 1.0 Aldrin 0.5 Endrin 1.0 Dieldrin 1.0 Dilute and re-analyze samples with one or more analytes at concentrations exceeding the range of the calibration curve. Results for such re-analyses should fall within the mid-range of the calibration curve. Report results and submit documentation for both analyses. Second column confirmation is required for all positive results. Perform confirmation analyses on a column of a phase different from that used for quantitation. Confirmation analyses must meet all instrument calibration criteria and blank acceptance criteria specified in Table 2, above. 8081CRF 6 of 5 Revision 12/03/1999 ------- |