United States
Environmental Protection
\r ^1 M^k. Agency
Office of Water
www.epa.gov	1990
Method 1624, Revision C:
Volatile Organic Compounds by
Isotope Dilution GCMS

-------
Method 1624
Revision C
Volatile Organic Compounds by Isotope Dilution
GCMS

-------
Method 1624
Volatile Organic Compounds by Isotope Dilution GCMS
1.	Scope and application
1.1	This method is designed to meet the survey requirements of the USEPA ITD. The method
is used to determine the volatile toxic organic pollutants associated with the Clean Water
Act (as amended 1987); the Resource Conservation and Recovery Act (as amended in
1986); the Comprehensive Environmental Response, Compensation, and Liability Act (as
amended in 1986); and other compounds amenable to purge and trap gas
chromatography/mass spectrometry (GCMS).
1.2	The chemical compounds listed in Tables 1 and 2 may be determined in waters, soils, and
municipal sludges by the method.
1.3	The detection limits of the method are usually dependent on the level of interferences
rather than instrumental limitations. The levels in Table 3 typify the minimum quantities
that can be detected with no interferences present.
1.4	The GCMS portions of the method are for use only by analysts experienced with GCMS
or under the close supervision of such qualified persons. Laboratories unfamiliar with
analysis of environmental samples by GCMS should run the performance tests in
Reference 1 before beginning.
2.	Summary of method
2.1	The percent solids content of the sample is determined. If the solids content is known or
determined to be less than 1%, stable isotopically labeled analogs of the compounds of
interest are added to a 5-mL sample and the sample is purged with an inert gas at 20 to
25°C in a chamber designed for soil or water samples. If the solids content is greater than
one, mL of reagent water and the labeled compounds are added to a 5-aliquot of sample
and the mixture is purged at 40°C. Compounds that will not purge at 20 to 25°C or at
40°C are purged at 75 to 85°C (see Table 2). In the purging process, the volatile
compounds are transferred from the aqueous phase into the gaseous phase where they
are passed into a sorbent column and trapped. After purging is completed, the trap is
backflushed and heated rapidly to desorb the compounds into a gas chromatograph (GC).
The compounds are separated by the GC and detected by a mass spectrometer (MS)
(References 2 and 3). The labeled compounds serve to correct the variability of the
analytical technique.
2.2	Identification of a pollutant (qualitative analysis) is performed in one of three ways: (1)
For compounds listed in Table 1 and other compounds for which authentic standards are
available, the GCMS system is calibrated and the mass spectrum and retention time for
each standard are stored in a user created library. A compound is identified when its
retention time and mass spectrum agree with the library retention time and spectrum. (2)
For compounds listed in Table 2 and other compounds for which standards are not
available, a compound is identified when the retention time and mass spectrum agree
with those specified in this method. (3) For chromatographic peaks which are not
identified by (1) and (2) above, the background corrected spectrum at the peak maximum
2

-------
is compared with spectra in the EPA/NIH mass spectral file (Reference 4). Tentative
identification is established when the spectrum agrees (see Section 12).
2.3	Quantitative analysis is performed in one of four ways by GCMS using extracted ion
current profile (EICP) areas: (1) For compounds listed in Table 1 and other compounds
for which standards and labeled analogs are available, the GCMS system iscalibrated and
the compound concentration is determined using an isotope dilution technique. (2) For
compounds listed in Table 1 and for other compounds for which authentic standards but
no labeled compounds are available, the GCMS system is calibrated and the compound
concentration is determined using an internal standard technique. (3) For compounds
listed in Table 2 and other compounds for which standards are not available, compound
concentrations are determined using known response factors. (4) For compounds for
which neither standards nor known response factors are available, compound
concentration is determined using the sum of the EICP areas relative to the sum of the
EICP areas of the nearest eluted internal standard.
2.4	The quality of the analysis is assured through reproducible calibration and testing of the
purge and trap and GCMS systems.
3

-------
Table 1. Volatile Organic Compounds Determined by GCMS Using Isotope Dilution and Internal
Standard Techniques
Pollutant	Labeled Compound


CAS
EPA



CAS
EPA
Compound
STORET
Registry
EGD
NPDES
Analog
Registry
EGD
Acetone
81552
67-64-1
516
V


d6
666-52-4
616
V
Acrolein
34210
107-02-8
002
V
001
V
d4
33984-05-3
202
V
Acrylonitrile
34215
107-13-1
003
V
002
V
d3
53807-26-4
203
V
Benzene
34030
71-43-2
004
V
003
V
d6
1076-43-3
204
V
Bromodichloromethane
32101
75-27-4
048
V
012
V
13C
93952-10-4
248
V
Bromoform
32104
75-25-2
047
V
005
V
13C
72802-81-4
247
V
Bromomethane
34413
74-83-9
046
V
020
V
d3
1111-88-2
246
V
Carbon tetrachloride
32102
56-23-5
006
V
006
V
13C
32488-50-9
206
V
Chlorobenzene
34301
108-90-7
007
V
007
V
d5
3114-55-4
207
V
Chloroethane
34311
75-00-3
016
V
009
V
d5
19199-91-8
216
V
2-Chloroethylvinyl ether
34576
110-75-8
019
V
010
V




Chloroform
32106
67-66-3
023
V
011
V
13C
31717-44-9
223
V
Chloromethane
34418
74-87-3
045
V
021
V
d3
1111-89-3
245
V
Dibromochloromethane
32105
124-48-1
051
V
008
V
13C
93951-99-6
251
V
1,1 -Dichloroethane
34496
75-34-3
013
V
014
V
d3
56912-77-7
213
V
1,2-Dichloroethane
32103
107-06-2
010
V
015
V
d4
17070-07-0
210
V
1,1 -Dichloroethene
34501
75-35-4
029
V
016
V
d2
22280-73-5
229
V
trans-1,2-Dichlorethene
34546
156-60-5
030
V
026
V
d3
42366-47-2
230
V
1,2-Dichloropropane
34541
78-87-5
032
V
017
V
d6
93952-08-0
232
V
trans-1,3-
34699
10061-02-6
033
V


d4
93951-86-1
233
V
Dichloropropene










Diethyl ether
81576
60-29-7
515
V


dio
2679-89-2
615
V
p-Dioxane
81582
123-91-1
527
V


d8
17647-74-4
627
V
Ethylbenzene
34371
100-41-4
038
V
019
V
dio
25837-05-2
238
V
Methylene chloride
34423
75-09-2
044
V
022
V
d2
1665-00-5
244
V
Methyl ethyl ketone
81595
78-93-3
514
V


d3
53389-26-7
614
V
1,1,2,2-
34516
79-34-5
015
V
023
V
d2
33685-54-0
215
V
T etrachlor oethane










T etrachlor oethene
34475
127-18-4
085
V
024
V
13c2
32488-49-6
285
V
Toluene
34010
108-88-3
086
V
025
V
d8
2037-26-5
286
V
1,1,1 -T richloroethane
34506
71-55-6
011
V
027
V
d3
2747-58-2
211
V
1,1,2-T richloroethane
34511
79-00-5
014
V
028
V
13C
93952-09-1
214
V
Trichloroethene
39180
79-01-6
087
V
029
V
13c2
93952-00-2
287
V
Vinyl chloride
39175
75-01-4
088
V
031
V
d3
6745-35-3
288
V
4

-------
Table 2. Volatile Organic Compounds to be Determined by Reverse Search and Quantitation
Using Known Retention Times, Response Factors, Reference Compounds, and Mass Spectra
EGD No.	Compound	CAS Registry
532
Allyl alcohol 1
107-18-6
533
Carbon disulfide
75-15-0
534
2-Chloro-1,3-butadiene (Chloroprene)
126-99-8
535
Chloroacetonitrile1
107-14-2
536
3-Chloropropene
107-05-1
537
Crotonaldehyde1
123-73-9
538
1,2-Dibromoethane (EDB)
106-93-3
539
Dibromomethane
74-95-3
540
trans-1,4-Dichloro-2-butene
110-57-6
541
1,3-Dichloropropane
142-28-9
542
cis-1,3-Dichloropropene
10061-01-5
543
Ethyl cyanide 1
107-12-0
544
Ethyl methacrylate
97-63-2
545
2-Hexanone
591-78-6
546
Iodomethane
74-88-4
547
Isobutyl alcohol1
78-83-1
548
Methacrylonitrile
126-98-7
549
Methyl methacrylate
78-83-1
550
4-Methyl-2-pentanone
108-10-1
551
1,1,1,2-T etrachloroethane
630-20-6
552
T richlorofluoromethane
75-69-4
553
1,2,3-Trichloropropane
96-18-4
554
Vinyl acetate
108-05-4
951
m-Xylene
108-38-3
952
o- and p-Xylene

1 Determined at a purge temperature of 75-85°C.
5

-------
3.
Contamination and interferences
3.1	Impurities in the purge gas, organic compounds out-gassing from the plumbing upstream
of the trap, and solvent vapors in the laboratory account for the majority of contamination
problems. The analytical system is demonstrated to be free from interferences under
conditions of the analysis by analyzing reagent water blanks initially and with each
sample batch (samples analyzed on the same 8-hour shift), as described in Section 8.5.
3.2	Samples can be contaminated by diffusion of volatile organic compounds (particularly
methylene chloride) through the bottle seal during shipment and storage. A field blank
prepared from reagent water and carried through the sampling and handling protocol
may serve as a check on such contamination.
3.3	Contamination by carry-over can occur when high level and low level samples are
analyzed sequentially. To reduce carry-over, the purging device (Figure 1 for samples
containing less than one percent solids; Figure 2 for samples containing one percent solids
or greater) is cleaned or replaced with a clean purging device after each sample is
analyzed. When an unusually concentrated sample is encountered, it is followed by
analysis of a reagent water blank to check for carry-over. Purging devices are cleaned by
washing with soap solution, rinsing with tap and distilled water, and drying in an oven
at 100 to 125°C. The trap and other parts of the system are also subject to contamination;
therefore, frequent bakeout and purging of the entire system may be required.
3.4	Interferences resulting from samples will vary considerably from source to source,
depending on the diversity of the site being sampled.
6

-------
Table 3. Gas Chromatography of Purgeable Organic Compounds
Method Detection
Retention Time	Limit4
Minimum Low High
EGD
No.1
Compound
Mean
(sec)
EGD
Ref
Relative 2
Level3
(ub/l)
Solids
(Ug/kg)
Solids
(Ug/kg)
245
Chloromethane-d3
147
181
0.141-0.270
50


345
Chloromethane
148
245
0.922-1.210
50
2077
13
246
Bromomenthane-d3
243
181
0.233-0.423
50


346
Bromomethane
246
246
0.898-1.195
50
00
11
288
Vinyl chloride-d3
301
181
0.286-0.501
50


388
Vinyl chloride
304
288
0.946-1.023
10
1907
11
216
Chloroethane-d5
378
181
0.373-0.620
50


316
Chloroethane
386
216
0.999-1.060
50
7897
24
244
Methylene chloride-d2
512
181
0.582-0.813
10


344
Methylene chloride
517
244
0.999-1.017
10
566 7
O
00
Cvl
546
Iodomethane
498
181
0.68



616
Acetone-d6
554
181
0.628-0.889
50


716
Acetone
565
616
0.984-1.019
7 50
3561
322 7
202
Acrolein-d4
564
181
0.641-0.903
5
50

302
Acrolein
566
202
0.984—1.0185
50
3777
18
203
Acrylonitrile-dj
606
181
0.735-0.926
50


303
Acrylonitrile
612
203
0.985-1.030
50
360 7
9
533
Carbon disulfide
631
181
0.86



552
T richlorofluoromethane
663
181
0.91



543
Ethyl cyanide
672
181
0.92



229
1,1 -Dichloroethene-d2
696
181
0.903-0.976
10


329
1,1 -Dichloroethene
696
229
0.999-1.011
10
31
5
536
3-Chloropropene
696
181
0.95



532
Allyl alcohol
703
181
0.96



181
Bromochloromethane (I.S.)
730
181
1.000-1.000
10


213
1,1 -Dichloroethane-dj
778
181
1.031-1.119
10


313
1,1 -Dichloroethane
786
213
0.999-1.014
10
16
1
615
Diethyl ether-d10
804
181
1.067-1.254
50


715
Diethyl ether
820
615
1.010-1.048
50
63
12
230
trans-l,2-Dichloroethene-d2
821
181
1.056-1.228
10


330
trans-1,2-Dichloroethene
821
230
0.996-1.011
10
41
3
614
Methyl ethyl ketone-d3
840
181
0.646-1.202
50


714
Methyl ethyl ketone
848
614
0.992-1.055
50
241 7
O
00
223
Chloroform-13Cj
861
181
1.092-1.322
10


323
Chloroform
861
223
0.961-1.009
10
21
2

-------
EGD
No.1
535
210
310
539
548
547
211
311
627
727
206
306
554
248
348
534
537
232
332
542
287
387
541
204
304
251
351
214
314
233
333
019
538
182
8
Method Detection
Retention Time	Limit4
Minimum Low High

Mean
EGD

Level3
Solids
Solids
Compound
(sec)
Ref
Relative 2
(Ufi/L)
(Ufi/ks)
(Ufi/ks)
Chloroacetonitrile
884
181
1.21



l,2-Dichloroethane-d4
901
181
1.187-1.416
10


1,2-Dichloroethane
910
210
0.973-1.032
10
23
3
Dibromomethane
910
181
1.25



Methacrylonitrile
921
181
1.26



Isobutyl alcohol
962
181
1.32



l,l,l-Trichloroethane-13C2
989
181
1.293-1.598
10


1,1,1 -T richloroethane
999
211
0.989-1.044
10
16
4
p-Dioxane-d8
982
181
1.262—1.4485
50


p-Dioxane
1001
627
1.008—1.0405
50
-
1407
Carbon tetrachloride-13C2
1018
182
0.754-0.805
10


Carbon tetrachloride
1018
206
0.938-1.005
10
87
9
Vinyl acetate
1031
182
0.79



Bromodichloromethane-13Cj
1045
182
0.766-0.825
10


Bromodichloromethane
1045
248
0.978-1.013
10
28
3
2-Chloro-1,3-butadiene
1084
182
0.83



Crotonaldehyde
1098
182
0.84



1,2-Dichloropropane-d6
1123
182
0.830-0.880
10


1,2-Dichloropropane
1134
232
0.984-1.018
10
29
5
cis-1,3-Dichloropropene
1138
182
0.87



Trichloroethene-13C2
1172
182
0.897-0.917
10


Trichloroethene
1187
287
0.991-1.037
10
41
2
1,3-Dichloropropane
1196
182
0.92



Benezene-d6
1200
182
0.888-0.952
10


Benezene
1212
204
1.002-1.026
10
23
8
Chlorodibromomethane-13Cj
1222
182
0.915-0.949
10


Chlorodibromomethane
1222
231
0.989-1.030
10
15
2
l,l,2-Trichloroethane-13C2
1224
182
0.922-0.953
10


1,1,2-T richloroethane
1224
214
0.975-1.027
10
26
1
trans-1,3-Dichloropropene-
cL
1226
182
0.922-0.959
10


trans-1,3-Dichloropropene
1226
233
0.993-1.016
10
	6,7
__6,7
2-Chloroethyvinyl ether
1278
182
0.983-1.026
10
122
21
1,2-Dibromoethane
1279
182
0.98



2-bromo-1 -chloropropane
1306
182
1.000-1.000
10



-------
Method Detection
Retention Time	Limit4
Minimum Low High
EGD
No.1
Compound
Mean
(sec)
EGD
Ref
Relative 2
Level3
(Ufi/L)
Solids Solids
(Ug/kg) (ug/kg)
549
Methyl methacrylate
1379
182
1.06


247
Bromoform-13Cj
1386
182
1.048-1.087
10

347
Bromoform
1386
247
0.992-1.003
10
91 7
551
1,1,1,2-T etrachloroethane
1408
182
1.08


550
4-Methyl-2-pentanone
1435
183
0.92


553
1,2,3-Trichloropropane
1520
183
0.98


215
l,l,2,2-Tetrachloroethane-d2
1525
183
0.969-0.996
10

315
1,1,2,2-T etrachloroethane
1525
215
0.890-1.016
10
20 6
545
2-Hexanone
1525
183
0.98


285
Tetrachloroethene-13C2
1528
183
0.966-0.996
10

385
T etrachloroethene
1528
285
0.997-1.003
10
106 10
540
trans-l,4-Dichloro-2-butene
1551
183
1.00


183
1,4-Dichlorobutane (int std)
1555
183
1.000-1.000
10

544
Ethyl methacrylate
1594
183
1.03


286
Toluene-d8
1603
183
1.016-1.054
10

386
Toluene
1619
286
1.001-1.019
10
27 4
207
Chlorobenzene-d5
1679
183
1.066-1.135
10

307
Chlorobenzene
1679
207
0.914-1.019
10
21 587
238
Ethylbenzene-d10
1802
183
1.144-1.293
10

338
Ethylbenzene
1820
238
0.981-1.018
10
28 4
185
Bromofluorobenzene
1985
183
1.255-1.290
10

951
m-Xylene
2348
183
1.51
10

952
o- and p-Xylene
2446
183
1.57
10

1
2
Reference numbers beginning with 0, 1, 5, or 9 indicate a pollutant quantified by the
internal standard method; reference numbers beginning with 2 or 6 indicate a labeled
compound quantified by the internal standard method; reference numbers beginning with
3 or 7 indicate a pollutant quantified by isotope dilution.
The retention time limits in this column are based on data from four wastewater
laboratories. The single values for retention times in this column are based on data from
one wastewater laboratory.
This is a minimum level at which the analytical system shall give recognizable mass
spectra (background corrected) and acceptable calibration points when calibrated using
reagent water. The concentration in the aqueous or solid phase is determined using the
equations in Section 13.
Method detection limits determined in digested sludge (low solids) and in filter cake or
compost (high solids).
Specification derived from related compound.

-------
6	An unknown interference in the particular sludge studied precluded measurement of the
method detection limit (MDL) for this compound.
7	Background levels of these compounds were present in the sludge resulting in higher than
expected MDLs. The MDL for these compounds is expected to be approximately 20
\ig/kg (100 to 200 \ig/kg for the gases and water soluble compounds) for the low solids
method and 5 to 10 \ig/kg (25 to 50 \ig/kg for the gases and water soluble compounds)
for the high solids methods, with no interferences present.
Column: 2.4 m (8 ft) x 2 mm I.D. glass, packed with 1% SP-1000 coated on 60/80 Carbopak B.
Carrier gas: Helium at 40 mL/min.
Temperature program: 3 min at 45°C, 8°C/min to 240°C, hold at 240°C for 15 minutes.
4.	Safety
4.1	The toxicity or carcinogenicity of each compound or reagent used in this method has not
been precisely determined; however, each chemical compound should be treated as a
potential health hazard.
Exposure to these compounds should be reduced to the lowest possible level. The
laboratory is responsible for maintaining a current awareness file of OSHA regulations
regarding the safe handling of the chemicals specified in this method. A reference file of
data handling sheets should also be made available to all personnel involved in these
analyses. Additional information on laboratory safety can be found in References 5
through 7.
4.2	The following compounds covered by this method have been tentatively classified as
known or suspected human or mammalian carcinogens: benzene, carbon tetrachloride,
chloroform, and vinyl chloride. Primary standards of these toxic compounds should be
prepared in a hood, and a NIOSH/MESA approved toxic gas respirator should be worn
when high concentrations are handled.
5.	Apparatus and materials
5.1	Sample bottles for discrete sampling.
5.1.1	Bottle: 25- to 40-mL with screw—cap (Pierce 13075, or equivalent). Detergent
—wash, rinse with tap and distilled water, and dry at >105°C for a minimum of
1 hour before use.
5.1.2	Septum: Teflon-faced silicone (Pierce 12722, or equivalent), cleaned as above and
baked at 100 to 200°C for 1 hour minimum.
5.2	Purge and trap device: Consists of purging device, trap, and desorber.
5.2.1 Purging devices for water and soil samples.
5.2.1.1 Purging device for water samples Designed to accept 5-mL samples
with water column at least 3 cm deep. The volume of the gaseous
head space between the water and trap shall be less than 15 mL. The
purge gas shall be introduced less than 5 mm from the base of the
water column and shall pass through the water as bubbles with a
10

-------
diameter less than 3 mm. The purging device shown in Figure 1 meets
these criteria.
5.2.1.2 Purging device for solid samples: Designed to accept 5 g of solids plus
5 mL of water. The volume of the gaseous head space between the
water and trap shall be less than 25 mL. The purge gas shall be
introduced less than 5 mm from the base of the sample and shall pass
through the water as bubbles with a diameter less than 3 mm. The
purging device shall be capable of operating at ambient temperature
(20 to 25°C) and of being controlled at temperatures of 40°C (±2°C) and
80°C (±5°C) while the sample is being purged. The purging device
shown in Figure 2 meets these criteria.
5.2.2	Trap: 25 to 30 cm long x 2.5 mm I.D. minimum, containing the following:
5.2.2.1	Methyl silicone packing: 1cm (±0.2cm), 3% OV-1 on 60/80 mesh
Chromosorb W, or equivalent.
5.2.2.2	Porous polymer: 15cm (±1.0 cm), Tenax GC (2,6-diphenylene oxide
polymer), 60/80 mesh, chromatographic grade, or equivalent.
5.2.2.3	Silica gel: 8cm (±1.0 cm), Davison Chemical, 35/60 mesh, grade 15, or
equivalent. The trap shown in Figure 3 meets these specifications.
5.2.3	Desorber: Shall heat the trap to 175°C (±5°C) in 45 seconds or less. The polymer
section of the trap shall not exceed a temperature of 180°C and the remaining
sections shall not exceed 220°C during desorb, and no portion of the trap shall
exceed 225°C during bakeout. The desorber shown in Figure 3 meets these
specifications.
5.2.4	The purge and trap device may be a separate unit, or coupled to a GC as shown
in Figures 4 and 5.
5.3	Gas chromatograph: Shall be linearly temperature programmable with initial and final
holds, shall contain a glass jet separator as the MS interface, and shall produce results
which meet the calibration (Section 7), quality assurance (Section 8), and performance tests
(Section 11) of this method.
5.3.1 Column: 2.8 • 0.4 m x 2 • 0.5 mm I.D. glass, packed with 1% SP-1000 on Carbopak
B, 60/80 mesh, or equivalent.
5.4	Mass spectrometer: 70 eV electron impact ionization; shall repetitively scan from 20 to
250 amu every 2 to 3 seconds, and produce a unit resolution (valleys between m/z 174
to 176 less than 10% of the height of the m/z 175 peak), background corrected mass
spectrum from 50 ng 4-bromofluorobenzene (BFB) injected into the GC. The BFB
spectrum shall meet the mass-intensity criteria in Table 4. All portions of the GC column,
transfer lines, and separator which connect the GC column to the ion source shall remain
at or above the column temperature during analysis to preclude condensation of less
volatile compounds.
5.5	Data system: Shall collect and record MS data, store mass-intensity data in spectral
libraries, process GCMS data and generate reports, and shall calculate and record response
factors.
11

-------

Table 4

BFB Mass-Intensity Specifications
m/z
Intensity Required
50
15 to 40% of m/z 95
75
30 to 60% of m/z 95
95
base peak, 100%
96
5 to 9% of m/z 95
173
less than 2% of m/z 174
174
greater than 50% of m/z 95
175
5 to 9% of m/z 174
176
95 to 101% of m/z 174
177
5 to 9% of m/z 176
5.5.1	Data acquisition: Mass spectra shall be collected continuously throughout the
analysis and stored on a mass storage device.
5.5.2	Mass spectral libraries: User-created libraries containing mass spectra obtained
from analysis of authentic standards shall be employed to reverse search GCMS
runs for the compounds of interest (Section 7.2).
5.5.3	Data processing: The data system shall be used to search, locate, identify, and
quantify the compounds of interest in each GCMS analysis. Software routines
shall be employed to compute retention times and EICP areas. Displays of
spectra, mass chromatograms, and library comparisons are required to verify
results.
5.5.4	Response factors and multipoint calibrations: The data system shall be used to
record and maintain lists of response factors (response ratios for isotope dilution)
and generate multi-point calibration curves (Section 7). Computations of relative
standard deviation (coefficient of variation) are useful for testing calibration
linearity. Statistics on initial and ongoing performance shall be maintained
(Sections 8 and 11).
5.6	Syringes: 5-mL glass hypodermic, with Luer-lok tips.
5.7	Micro syringes: 10-, 25-, and 100 |iL.
5.8	Syringe valves: 2-way, with Luer ends (Teflon or Kel-F).
5.9	Syringe: 5-mL, gas-tight, with shut-off valve.
5.10	Bottles: 15-mL, screw-cap with Teflon liner.
5.11	Balances.
5.11.1	Analytical, capable of weighing 0.1 mg.
5.11.2	Top-loading, capable of weighing 10 mg.
5.12	Equipment for determining percent moisture.
12

-------
5.12.1	Oven,capable of being temperature-controlled at 110°C (±5°C).
5.12.2	Dessicator.
5.12.3	Beakers: 50 to 100-mL.
6. Reagents and standards
6.1	Reagent water: Water in which the compounds of interest and interfering compounds are
not detected by this method (Section 11.7). It may be generated by any of the following
methods:
6.1.1	Activated carbon: pass tap water through a carbon bed (Calgon Filtrasorb-300,
or equivalent).
6.1.2	Water purifier: Pass tap water through a purifier (Millipore Super Q, or
equivalent).
6.1.3	Boil and purge: Heat tap water to between 90 and 100°C and bubble contaminant
free inert gas through it for approximately 1 hour. While still hot, transfer the
water to screw-cap bottles and seal with a Teflon-lined cap.
6.2	Sodium thiosulfate: ACS granular.
6.3	Methanol: Pesticide-quality or equivalent.
6.4	Standard solutions: Purchased as solutions or mixtures with certification to their purity,
concentration, and authenticity, or prepared from materials of known purity and
composition. If compound purity is 96% or greater, the weight may be used without
correction to calculate the concentration of the standard.
6.5	Preparation of stock solutions: Prepare in methanol using liquid or gaseous standards per
the steps below. Observe the safety precautions given in Section 4.
6.5.1	Place approximately 9.8 mL of methanol in a 10-mL ground-glass-stoppered
volumetric flask. Allow the flask to stand unstoppered for approximately 10
minutes or until all methanol wetted surfaces have dried. In each case, weigh the
flask, immediately add the compound, then immediately reweigh to prevent
evaporation losses from affecting the measurement.
6.5.1.1	Liquids: Using a 100 |iL syringe, permit 2 drops of liquid to fall into
the methanol without contacting the neck of the flask. Alternatively,
inject a known volume of the compound into the methanol in the flask
using a micro-syringe.
6.5.1.2	Gases (chloromethane, bromomethane, chloroethane, vinyl chloride):
Fill a valved 5-mL gas-tight syringe with the compound. Lower the
needle to approximately 5 mm above the methanol meniscus. Slowly
introduce the compound above the surface of the meniscus. The gas
will dissolve rapidly in the methanol.
6.5.2	Fill the flask to volume, stopper, then mix by inverting several times. Calculate
the concentration in mg/mL ((Jg/ pL) from the weight gain (or density if a known
volume was injected).
13

-------
6.5.3	Transfer the stock solution to a Teflon-sealed screw-cap bottle. Store, with
minimal headspace, in the dark at -10 to -20°C.
6.5.4	Prepare fresh standards weekly for the gases and 2-chloroethylvinyl ether. All
other standards are replaced after one month, or sooner if comparison with check
standards indicate a change in concentration. Quality control check standards
that can be used to determine the accuracy of calibration standards are available
from the US Environmental Protection Agency, Environmental Monitoring and
Support Laboratory, Cincinnati, Ohio.
6.6	Labeled compound spiking solution: From stock standard solutions prepared as above,
or from mixtures, prepare the spiking solution to contain a concentration such that a 5-
to 10- |iL spike into each 5-mL sample, blank, or aqueous standard analyzed will result
in a concentration of 20 ug/L of each labeled compound. For the gases and for the water
soluble compounds (acrolein, acrylonitrile, acetone, diethyl ether, p-dioxane, and MEK),
a concentration of 100 ug/L may be used. Include the internal standards (Section 7.5) in
this solution so that a concentration of 20 ug/L in each sample, blank, or aqueous
standard will be produced.
6.7	Secondary standards: Using stock solutions, prepare a secondary standard in methanol
to contain each pollutant at a concentration of 500 |ig/mL. For the gases and water
soluble compounds (Section 6.6), a concentration of 2.5 mg/mL may be used.
6.7.1	Aqueous calibration standards: Using a 25-|iL syringe, add 20 |iL of the
secondary standard (Section 6.7) to 50, 100, 200, 500, and 1000 mL of reagent
water to produce concentrations of 200, 100, 50, 20, and 10 Ug/L, respectively.
If the higher concentration standard for the gases and water soluble compounds
was chosen (Section 6.6), these compounds will be at concentrations of 1000, 500,
250, 100, and 50 Ug/L in the aqueous calibration standards.
6.7.2	Aqueous performance standard: An aqueous standard containing all pollutants,
internal standards, labeled compounds, and BFB is prepared daily, and analyzed
each shift to demonstrate performance (Section 11). This standard shall contain
either 20 or 100 Ug/L of the labeled and pollutant gases and water soluble
compounds, 10 Ug/L BFB, and 20 Ug/L of all other pollutants, labeled
compounds, and internal standards. It may be the nominal 20 Ug/L aqueous
calibration standard (Section 6.7.1).
6.7.3	A methanolic standard containing all pollutants and internal standards is
prepared to demonstrate recovery of these compounds when syringe injection and
purge-and-trap analyses are compared. This standard shall contain either 100
|ig/mL or 500 y\g/mL of the gases and water soluble compounds, and 100 |ig/mL
of the remaining pollutants and internal standards (consistent with the amounts
in the aqueous performance standard in 6.7.2).
6.7.4	Other standards which may be needed are those for test of BFB performance
(Section 7.1) and for collection of mass spectra for storage in spectral libraries
(Section 7.2).
7. Calibration
14

-------
Calibration of the GCMS system is performed by purging the compounds of interest and their
labeled analogs from reagent water at the temperature to be used for analysis of samples.
7.1	Assemble the gas chromatographic apparatus and establish operating conditions given in
Table 3. By injecting standards into the GC, demonstrate that the analytical system meets
the minimum levels in Table 3 for the compounds for which calibration is to be
performed, and the mass-intensity criteria in Table 4 for 50 ng BFB.
7.2	Mass spectral libraries: Detection and identification of the compounds of interest are
dependent upon the spectra stored in user created libraries.
7.2.1	For the compounds in Table 1 and other compounds for which the GCMS is to
be calibrated, obtain a mass spectrum of each pollutant and labeled compound
and each internal standard by analyzing an authentic standard either singly or as
part of a mixture in which there is no interference between closely eluted
components. Examine the spectrum to determine that only a single compound
is present. Fragments not attributable to the compound under study indicate the
presence of an interfering compound. Adjust the analytical conditions and scan
rate (for this test only) to produce an undistorted spectrum at the GC peak
maximum. An undistorted spectrum will usually be obtained if five complete
spectra are collected across the upper half of the GC peak. Software algorithms
designed to "enhance" the spectrum may eliminate distortion, but may also
eliminate authentic m/z's or introduce other distortion.
7.2.2	The authentic reference spectrum is obtained under BFB tuning conditions
(Section 7.1 and Table 4) to normalize it to spectra from other instruments.
7.2.3	The spectrum is edited by saving the five most intense mass spectral peaks and
all other mass spectral peaks greater than 10% of the base peak. The spectrum
may be further edited to remove common interfering masses. If five mass
spectral peaks cannot be obtained under the scan conditions given in Section 5.4,
the mass spectrometer may be scanned to an m/z lower than 20 to gain
additional spectral information. The spectrum obtained is stored for reverse
search and for compound confirmation.
7.2.4	For the compounds in Table 2 and other compounds for which the mass spectra,
quantitation m/z's, and retention times are known but the instrument is not to
be calibrated, add the retention time and reference compound (Table 3); the
response factor and the quantitation m/z (Table 5); and spectrum (Appendix A)
to the reverse search library. Edit the spectrum per Section 7.2.3, if necessary.
7.3	Assemble the purge-and-trap device. Pack the trap as shown in Figure 3 and condition
overnight at 170 to 180°C by backflushing with an inert gas at a flow rate of 20 to 30
mL/min. Condition traps daily for a minimum of 10 minutes prior to use.
7.3.1 Analyze the aqueous performance standard (Section 6.7.2) according to the
purge-and-trap procedure in Section 10. Compute the area at the primary m/z
(Table 5) for each compound. Compare these areas to those obtained by injecting
1 |iL of the methanolic standard (Section 6.7.3) to determine compound recovery.
The recovery shall be greater than 20% for the water soluble compounds (Section
6.6), and 60 to 110% for all other compounds. This recovery is demonstrated
initially for each purge-and-trap GCMS system. The test is repeated only if the
15

-------
purge-and-trap or GCMS systems are modified in any way that might result in
a change in recovery.
7.3.2 Demonstrate that 100 ng toluene (or toluene-dg) produces an area at m^z 91 (or
99) approximately one-tenth that required to exceed the linear range of the
system. The exact value must be determined by experience for each instrument.
It is used to match the calibration range of the instrument to the analytical range
and detection limits required.
Table 5. Volatile Organic Compound Characteristic M/Z'S




Response purge

labeled
Primary
Reference
temp. Of:
Compound
Analog
m/z!
Compound2
O
0
O
00
o
0
O
CM
Acetone
d6
58/64


Acrolein
d4
56/60


Acrylonitrile
d3
53/56


Allyl alcohol

57
181
-3 0.20
Benzene
d6
78/84


2-Bromo-1 -chloropropane

77


Bromochloromethane4

128


Bromodichloromethane
13c
83/86


Bromoform
13c
173/176


Bromomethane
d3
96/99


Carbon disulfide

76
181
1.93 2.02
Carbon tetrachloride
13c
47/48


2-Chloro-1,3-butadiene

53
182
0.29 0.50
Chloroacetonitrile

75
181
-3 1.12
Chlorobenzene
d5
112/117


Chloroethane
d5
64/71


2-Chloroethylvinyl ether
d7
106/113


Chloroform
13c
85/86


Chloromethane
d3
50/53


3-Chloropropene

76
181
0.43 0.63
Crotonaldehyde

70
182
-3 0.090
Dibromochloromethane
13c
129/130


1,2-Dibromoethane

107
182
0.86 0.68
Dibromomethane

93
181
1.35 1.91
1,4-Dichlorobutane

55


trans-1,4-Dichloro-2-bu-

75
183
0.093 0.014
tene




1,1 -Dichloroethane
d3
63/66


16

-------
Response purge
labeled Primary Reference	temp. Of:
Compound
Analog
m/z!
Compound2
20 °C
OO
o
0
O
1,2-Dichloroethane
d4
62/67



1,1 -Dichloroethene
d2
61/65



trans-1,2-Dichlorethene
d2
61/65



1,2-Dichloropropane
d6
63/67



1,3-Dichloropropane

76
182
0.89
0.88
cis-1,3-Dichloropropene

75
182
0.29
0.41
trans-1,3-Dichloropropene
d4
75/79



Diethyl ether
dio
74/84



p-Dioxane
d8
88/96



Ethyl cyanide

54
181
(3)
1.26
Ethyl methacrylate

69
183
0.69
0.52
Ethylbenzene
dio
106/116



2-Hexanone

58
183
0.076
0.33
Iodomethane

142
181
4.55
2.55
Isobutyl alcohol

74
181
(3)
0.22
Methylene chloride
d2
84/88



Methyl ethyl ketone
d8
72/80



Methyl methacrylate

69
182
0.23
0.79
4-Methyl-2-pentanone

58
183
0.15
0.29
Methacrylonitrile

67
181
0.25
0.79
1,1,1,2-T etrachloroethane

131
182
0.20
0.25
1,1,2,2-T etrachloroethane
d2
83/84



T etrachloroethene
13c2
164/172



Toluene
d8
92/100



1,1,1 -Trichloroethane
d3
97/102



1,1,2-Trichloroethane
13c2
83/84



Trichloroethene
13c2
95/136



T richlorofluoromethane

101
181
2.31
2.19
1,2,3-Trichloropropane

75
183
0.89
0.72
Vinyl acetate

86
182
0.054
0.19
Vinyl chloride
d3
62/65



m-Xylene

106
183
1.69
-
0- and p-Xylene

106
183
3.33
-
181	= bromochloromethane
182	= 2-bromo-l-chloropropane
183	= 1,4-dichlorobutane
Not detected at a purge temperature of 20°C
Internal standard

-------
Note: Because the composition and purity of commercially-supplied isotopically labeled stan-
dard's may vary, the primary m/z of the labeled analogs given in this table should be used as
guidance. The appropriate m/z of the labeled analogs should be determined prior to use for
sample analysis. Deviations from the m/z's listed here must be documented by the laboratory
and submitted with the data.
7.4 Calibration by isotope dilution: The isotope dilution approach is used for the purgeable
organic compounds when appropriate labeled compounds are available and when
interferences do not preclude the analysis. If labeled compounds are not available, or
interferences are present, the internal standard method (Section 7.5) is used. A calibration
curve encompassing the concentration range of interest is prepared for each compound
determined. The relative response (RR) vs. concentration (l-ig/L) is plotted or computed
using a linear regression. An example of a calibration curve for toluene using toluene-d8
is given in Figure 6. Also shown are the ±10% error limits (dotted lines). Relative
response is determined according to the procedures described below. A minimum of five
data points are required for calibration (Section 7.4.4).
7.4.1	The relative response (RR) of pollutant to labeled compound is determined from
isotope ratio values calculated from acquired data. Three isotope ratios are used
in this process:
Rx = the isotope ratio measured in the pure pollutant (Figure 7A).
Ry = the isotope ratio of pure labeled compound (Figure 7B).
Rm = the isotope ratio measured in the analytical mixture of the
pollutant and labeled compounds (Figure 7 C.)
The correct way to calculate RR is:
_ (V (*,+ 1)
(*m - RX) (Ry + D
If Rm is not between 2Ry and 0.55 > the method does not apply and the sample
is analyzed by the internal standard method (Section 7.5).
7.4.2	In most cases, the retention times of the pollutant and labeled compound are the
same, and isotope ratios (R's) can be calculated from the EICP areas, where:
(area at mJz)
R = 		—
(area at m2/z)
If either of the areas is zero, it is assigned a value of one in the calculations; that
is, if: area of m/z = 50721,
area of mz/Z = 0,
then R = 50721/1 = 50720
The data from these analyses are reported to three significant figures (see Section
13.6). In order to prevent rounding errors from affecting the values to be
18

-------
reported, all calculations performed prior to the final determination of
concentrations should be carried out using at least four significant figures.
Therefore, the calculation of R above is rounded to four significant figures.
The m/z's are always selected such that Rx > Ry. When there is a difference in
retention times (RT) between the pollutant and labeled compounds, special
precautions are required to determine the isotope ratios.
Rx, Ry, and Rmare defined as follows:
[area m^z (at
Rx"

R =

1

1
[area
m2/z (at RT2)]
[area
mx!z (at
[area mJz (at RT2)]
7.4.3 An example of the above calculations can be taken from the data plotted in Figure
7 for toluene and toluene-d8. For these data:
= 168920 = 168900
1
IT = —— = 0.00001640
y 60960
R = 96868 - 1.171
m 82508
The RR for the above data is then calculated using the equation given in Section
7.4.1. For the example, rounded to four significant figures, RR = 1.174. Not all
labeled compounds elute before their pollutant analogs.
7.4.4	To calibrate the analytical system by isotope dilution, analyze a 5-mL aliquot of
each of the aqueous calibration standards (Section 6.7.1) spiked with an
appropriate constant amount of the labeled compound spiking solution (Section
6.6), using the purge-and-trap procedure in Section 10. Compute the RR at each
concentration.
7.4.5	Linearity: If the ratio of relative response to concentration for any compound is
constant (less than 20% coefficient of variation) over the five point calibration
range, an averaged relative response/concentration ratio may be used for that
compound; otherwise, the complete calibration curve for that compound shall be
used over the five point calibration range.
Calibration by internal standard: Used when criteria for isotope dilution (Section 7.4)
cannot be met. The method is applied to pollutants having no labeled analog and to the
labeled compounds. The internal standards used for volatiles analyses are
bromochloromethane, 2-bromo-l-chloropropane, and 1,4-dichlorobutane. Concentrations
19

-------
of the labeled compounds and pollutants without labeled analogs are computed relative
to the nearest eluting internal standard, as shown in Tables 3 and 5.
7.5.1	Response factors: Calibration requires the determination of response factors (RF)
which are defined by the following equation:
R (As * CJ,
(a,-s * Cs)
Where:
A = is the EICP area at the characteristic m/z for the compound in the daily standard.
Ais = is the EICP area at the characteristic m/z for the internal standard.
Cjs = is the concentration ([ig/L) of the internal standard.
Cs = is the concentration of the pollutant in the daily standard.
7.5.2	The response factor is determined at 10, 20, 50, 100, and 200 ]ig/L for the
pollutants (optionally at five times these concentrations for gases and water
soluble pollutants; see Section 6.7), in a way analogous to that for calibration by
isotope dilution (Section 7.4.4). The RF is plotted against concentration for each
compound in the standard (CJ to produce a calibration curve.
7.5.3	Linearity: If the response factor (RF) for any compound is constant (less than 35%
coefficient of variation) over the five-point calibration range, an averaged response
factor may be used for that compound; otherwise, the complete calibration curve
for that compound shall be used over the five-point range.
7.6	Combined calibration: By adding the isotopically labeled compounds and internal
standards (Section 6.6) to the aqueous calibration standards (Section 6.7.1), a single set of
analyses can be used to produce calibration curves for the isotope dilution and internal
standard methods. These curves are verified each shift (Section 11.5) by purging the
aqueous performance standard (Section 6.7.2). Recalibration is required only if calibration
and ongoing performance (Section 11.5) criteria cannot be met.
7.7	Elevated purge temperature calibration: Samples containing greater than 1% solids are
analyzed at a temperature of 40°C (±2°C) (Section 10). For these samples, the analytical
system may be calibrated using a purge temperature of 40°C(±2°C) in order to more
closely approximate the behavior of the compounds of interest in high solids samples.
8. Quality assurance/quality control
8.1 Each laboratory that uses this method is required to operate a formal quality assurance
program (Reference 8). The minimum requirements of this program consist of an initial
demonstration of laboratory capability, analysis of samples spiked with labeled
compounds to evaluate and document data quality, and analysis of standards and blanks
as tests of continued performance. Laboratory performance is compared to established
performance criteria to determine if the results of analyses meet the performance
characteristics of the method.
8.1.1 The analyst shall make an initial demonstration of the ability to generate
acceptable accuracy and precision with this method. This ability is established as
described in Section 8.2.
20

-------
8.1.2	The analyst is permitted to modify this method to improve separations or lower
the costs of measurements, provided all performance specifications are met. Each
time a modification is made to the method, the analyst is required to repeat the
procedure in Section 8.2 to demonstrate method performance.
8.1.3	Analyses of blanks are required to demonstrate freedom from contamination and
that the compounds of interest and interfering compounds have not been carried
over from a previous analysis (Section 3). The procedures and criteria for analysis
of a blank are described in Section 8.5.
8.1.4	The laboratory shall spike all samples with labeled compounds to monitor
method performance. This test is described in Section 8.3. When results of these
spikes indicate atypical method performance for samples, the samples are diluted
to bring method performance within acceptable limits (Section 14.2).
8.1.5	The laboratory shall, on an ongoing basis, demonstrate through the analysis of the
aqueous performance standard (Section 6.7.2) that the analysis system is in
control. This procedure is described in Sections 11.1 and 11.5.
8.1.6	The laboratory shall maintain records to define the quality of data that is
generated. Development of accuracy statements is described in Sections 8.4 and
11.5.2.
21

-------
Table 6. Acceptance Criteria for Performance Tests
Acceptance criteria at 20 Hg/L or as noted




Labeled



Labeled and native
compound
Labeled and


compound initial
recovery
native compound
EGD

precision and ac-
(Sect. 8.3 and
ongoing accuracy


curacy (Sect
. 8.2.3)
14.2)
(Sect. 11.5)
No.1
Compound
s ( ug/L)
X (ug/L)
P (%)
R (ug/L)
516
acetone*
51.0
77 - 153
35 - 165
55 - 145
002
acrolein*
72.0
32 - 168
37 - 163
7 - 190
003
acrylonitrile*
16.0
70 - 132
ns - 204
58 - 144
004
benzene
9.0
13 - 28
ns - 196
4 - 33
048
bromodichloro-
8.2
7 - 32
ns - 199
4 - 34

methane




047
bromoform
7.0
7 - 35
ns - 214
6 - 36
046
bromomethane
25.0
d - 54
ns - 414
d - 61
006
carbon
6.9
16 - 25
42 - 165
12 - 30

tetrachloride




007
chlorobenzene
8.2
14 - 30
ns - 205
4 - 35
016
chloroethane
15.0
d - 47
ns - 308
d - 51
019
2-chloroethylvinyl
36.0
d - 70
ns - 554
d - 79

ether




023
chloroform
7.9
12 - 26
18 - 172
8 - 30
045
chloromethane
26.0
d - 56
ns - 410
d - 64
051
dibromochloro-
7.9
11 - 29
16 - 185
8 - 32

methane




013
1,1 -dichloroethane
6.7
11 - 31
23 - 191
9 - 33
010
1,2-dichloroethane
7.7
12 - 30
12 - 192
8 - 33
029
1,1 -dichloroethene
12.0
d - 50
ns - 315
d - 52
030
trans-1,2-dichloro-
7.4
11 - 32
15 - 195
8 - 34

ethene




032
1,2-dichloropropane
19.0
d - 47
ns - 343
d - 51
033
trans-1,3-dichloro-
15.0
d - 40
ns - 284
d - 44

propene




515
diethyl ether*
44.0
75 - 146
44 - 156
55 - 145

-------
Acceptance criteria at 20 Hg/L or as noted
EGD
No.
Compound
Labeled and native
compound initial
precision and ac-
curacy (Sect. 8.2.3)
s ( ug/L) X (ug/L)
Labeled
compound
recovery
(Sect. 8.3 and
14.2)
Labeled and
native compound
ongoing accuracy
(Sect. 11.5)
R (ua/L)
527
p-dioxane*
7.2
13 - 27
ns - 239
11 - 29
038
ethylbenzene
9.6
16 - 29
ns - 203
5 - 35
044
methylene chloride
9.7
d - 50
ns - 316
d - 50
514
methyl ethyl
ketone*
57.0
66 - 159
36 - 164
42 - 158
015
1,1,2,2 -tetr achlor o-
ethane
9.6
11 - 30
5 - 199
7 - 34
085
tetrachloroethane
6.6
15 - 29
31 - 181
11 - 32
086
toluene
6.3
15 - 29
4 - 193
6 - 33
011
1,1,1-
trichloroethane
5.9
11 - 33
12 - 200
8 - 35
014
1,1,2-
trichloroethane
7.1
12 - 30
21 - 184
9 - 32
087
trichloroethene
8.9
17 - 30
35 - 196
12 - 34
088
vinyl chloride
28.0
d - 59
ns - 452
d - 65
* acceptance criteria at 100 Ug/L
d = detected; result must be greater than zero.
ns = no specification; limit would be below detection limit.
1 Reference numbers beginning with 0, 1, or 5 indicate a pollutant quantified by the internal
standard method; reference numbers beginning with 2 or 6 indicate a labeled compound
quantified by the internal Standard method; reference numbers beginning with 3 or 7
indicate a pollutant quantified by isotope dilution.
23

-------
8.2 Initial precision and accuracy: To establish the ability to generate acceptable precision and
accuracy, the analyst shall perform the following operations for compounds to be
calibrated:
8.2.1	Analyze two sets of four 5-mL aliquots (8 aliquots total) of the aqueous
performance standard (Section 6.7.2) according to the method beginning in Section
10.
8.2.2	Using results of the first set of four analyses in Section 8.2.1, compute the average
recovery (X) in Hg/L and the standard deviation of the recovery (s) in Hg/L for
each compound, by isotope dilution for pollutants with a labeled analog, and by
internal standard for labeled compounds and pollutants with no labeled analog.
8.2.3	For each compound, compare s and X with the corresponding limits for initial
precision and accuracy found in Table 6. If s and X for all compounds meet the
acceptance criteria, system performance is acceptable and analysis of blanks and
samples may begin. If, however, any individual s exceeds the precision limit or
any individual X falls outside the range for accuracy, system performance is
unacceptable for that compound.
NOTE: The large number of compounds in Table 6 present a substantial probability
that one or more will fail one of the acceptance criteria when all compounds are
analyzed. To determine if the analytical system is out of control, or if the
failure can be attributed to probability, proceed as follows:
8.2.4 Using the results of the second set of four analyses, compute s and X for only
those compounds which failed the test of the first set of four analyses (Section
8.2.3). If these compounds now pass, system performance is acceptable for all
compounds and analysis of blanks and samples may begin. If, however, any of
the same compounds fail again, the analysis system is not performing properly
for the compound (s) in question. In this event, correct the problem and repeat
the entire test (Section 8.2.1).
8.3	The laboratory shall spike all samples with labeled compounds to assess method
performance on the sample matrix.
8.3.1	Spike and analyze each sample according to the method beginning in Section 10.
8.3.2	Compute the percent recovery (P) of the labeled compounds using the internal
standard method (Section 7.5).
8.3.3	Compare the percent recovery for each compound with the corresponding labeled
compound recovery limit in Table 6. If the recovery of any compound falls
outside its warning limit, method performance is unacceptable for that compound
in that sample. Therefore, the sample matrix is complex and the sample is to be
diluted and reanalyzed, per Section 14.2.
8.4	As part of the QA program for the laboratory, method accuracy for wastewater samples
shall be assessed and records shall be maintained. After the analysis of five wastewater
samples for which the labeled compounds pass the tests in Section 8.3.3, compute the
24

-------
average percent recovery (P) and the standard deviation of the percent recovery (sp) for
the labeled compounds only. Express the accuracy assessment as a percent recovery
interval from P - 2sp to P + 2sp. For example, if P = 90% and sp = 10%, the accuracy
interval is expressed as 70 to 110%. Update the accuracy assessment for each compound
on a regular basis (e.g., after each 5 to 10 new accuracy measurements).
8.5	Blanks: Reagent water blanks are analyzed to demonstrate freedom from carry-over
(Section 3) and contamination.
8.5.1	The level at which the purge and trap system will carry greater than 5 Hg/L of
a pollutant of interest (Tables 1 and 2) into a succeeding blank shall be
determined by analyzing successively larger concentrations of these compounds.
When a sample contains this concentration or more, a blank shall be analyzed
immediately following this sample to demonstrate no carry-over at the 5 Hg/L
level.
8.5.2	With each sample lot (samples analyzed on the same 8-hour shift), a blank shall
be analyzed immediately after analysis of the aqueous performance standard
(Section 11.1) to demonstrate freedom from contamination. If any of the
compounds of interest (Tables 1 and 2) or any potentially interfering compound
is found in a blank at greater than 10 Hg/L (assuming a response factor of 1
relative to the nearest eluted internal standard for compounds not listed in Tables
1 and 2), analysis of samples is halted until the source of contamination is
eliminated and a blank shows no evidence of contamination at this level.
8.6	The specifications contained in this method can be met if the apparatus used is calibrated
properly, then maintained in a calibrated state. The standards used for calibration
(Section 7), calibration verification (Section 11.5) and for initial (Section 8.2) and ongoing
(Section 11.5) precision and accuracy should be identical, so that the most precise results
will be obtained. The GCMS instrument in particular will provide the most reproducible
results if dedicated to the settings and conditions required for the analyses of volatiles by
this method.
8.7	Depending on specific program requirements, field replicates may be collected to
determine the precision of the sampling technique, and spiked samples may be required
to determine the accuracy of the analysis when the internal method is used.
9. Sample collection, preservation, and handling
9.1	Grab samples are collected in glass containers having a total volume greater than 20 mL.
For aqueous samples which pour freely, fill sample bottles so that no air bubbles pass
through the sample as the bottle is filled and seal each bottle so that no air bubbles are
entrapped. Maintain the hermetic seal on the sample bottle until time of analysis.
9.2	Samples are maintained at 0 to 4°C from the time of collection until analysis. If an
aqueous sample contains residual chlorine, add sodium thiosulfate preservative (10 mg/40
mL) to the empty sample bottles just prior to shipment to the sample site. EPA Methods
330.4 and 330.5 may be used for measurement of residual chlorine (Reference 9). If
preservative has been added, shake the bottle vigorously for one minute immediately after
filling.
9.3	For aqueous samples, experimental evidence indicates that some aromatic compounds,
notably benzene, toluene, and ethyl benzene are susceptible to rapid biological
25

-------
degradation under certain environmental conditions. Refrigeration alone may not be
adequate to preserve these compounds in wastewaters for more than seven days. For this
reason, a separate sample should be collected, acidified, and analyzed when these
aromatics are to be determined. Collect about 500 mL of sample in a clean container.
Adjust the pH of the sample to about 2 by adding HC1 (1 + 1) while stirring. Check pH
with narrow range (1.4 to 2.8) pH paper. Fill a sample container as described in Section
9.1. If residual chlorine is present, add sodium thiosulfate to a separate sample container
and fill as in Section 9.1.
9.4 All samples shall be analyzed within 14 days of collection.
10. Purge, trap, and GCMS analysis
Samples containing less than one percent solids are analyzed directly as aqueous samples
(Section 10.4). Samples containing one percent solids or greater are analyzed as solid
samples utilizing one of two methods, depending on the levels of pollutants in the
sample. Samples containing one percent solids or greater and low to moderate levels of
pollutants are analyzed by purging a known weight of sample added to 5 mL of reagent
water (Section 10.5). Samples containing 1% solids or greater and high levels of pollutants
are extracted with methanol, and an aliquot of the methanol extract is added to reagent
water and purged (Section 10.6).
10.1 Determination of percent solids.
10.1.1	Weigh 5 to 10 g of sample into a tared beaker.
10.1.2	Dry overnight (12 hours minimum) at 110°C (±5°C), and cool in a dessicator.
10.1.3	Determine percent solids as follows:
% solids = weight of sample dryx 100
weight of sample wet
10.2	Remove standards and samples from cold storage and bring to 20 to 25°C.
10.3	Adjust the purge gas flow rate to 40 (±4mL/min).
10.4	Samples containing less than 1% solids.
10.4.1	Mix the sample by shaking vigorously. Remove the plunger from a 5-mL syringe
and attach a closed syringe valve. Open the sample bottle and carefully pour the
sample into the syringe barrel until it overflows. Replace the plunger and
compress the sample. Open the syringe valve and vent any residual air while
adjusting the sample volume to 5 mL (±0.1 mL). Because this process of taking
an aliquot destroys the validity of the sample for future analysis, fill a second
syringe at this time to protect against possible loss of data.
10.4.2	Add an appropriate amount of the labeled compound spiking solution (Section
6.6) through the valve bore, then close the valve.
10.4.3	Attach the syringe valve assembly to the syringe valve on the purging device.
Open both syringe valves and inject the sample into the purging chamber. Purge
the sample per Section 10.7.
26

-------
10.5	Samples containing 1% solids or greater and low to moderate levels of pollutants.
10.5.1	Mix the sample thoroughly using a clean spatula.
10.5.2	Weigh 5 g (±1 g) of sample into a purging vessel (Figure 2). Record the weight
to three significant figures.
10.5.3	Add 5 mL (±0.1 mL) of reagent water to the vessel.
10.5.4	Using a metal spatula, break up any lumps of sample to disperse the sample in
the water.
10.5.5	Add an appropriate amount of the labeled compound spiking solution (Section
6.6) to the sample in the purge vessel. Place a cap on the purging vessel and and
shake vigorously to further disperse the sample. Attach the purge vessel to the
purging device, and purge the sample per Section 10.7.
10.6	Samples containing 1% solids or greater and high levels of pollutants, or samples
requiring dilution by a factor of more than 100 (see Section 13.4).
10.6.1	Mix the sample thoroughly using a clean spatula.
10.6.2	Weigh 5g (±1 g) of sample into a calibrated 15- to 25-mL centrifuge tube. Record
the weight of the sample to three significant figures.
10.6.3	Add 10 mL of methanol to the centrifuge tube. Cap the tube and shake it
vigorously for 15 to 20 seconds to disperse the sample in the methanol. Allow
the sample to settle in the tube. If necessary, centrifuge the sample to settle
suspended particles.
10.6.4	Remove approximately 0.1% of the volume of the supernatant methanol using a
15- to 25- |iL syringe. This volume will be in the range of 10 to 15 |iL.
10.6.5	Add this volume of the methanol extract to 5 mL reagent water in a 5 mL syringe,
and analyze per Section 10.4.1.
10.6.6	For further dilutions, dilute 1 mL of the supernatant methanol (Section 10.6.4) to
10 mL, 100 mL, 1000 mL, etc., in reagent water. Remove a volume of this
methanol extract/reagent water mixture equivalent to the volume in Section
10.6.4, add it to 5 mL reagent water in a 5 mL syringe, and analyze per Section
10.4.1.
10.7	Purge the sample for 11 minutes (±0.1 minute) at 20 to 25°C for samples containing less
than 1% solids. Purge samples containing one percent solids or greater at 40°(±2°). If the
compounds in Table 2 that do not purge at 20 to 40°C are to be determined, a purge
temperature of 80°C (±5°C) is used.
10.8	After the 11 minute purge time, attach the trap to the chromatograph and set the purge-
and- trap apparatus to the desorb mode (Figure 5). Desorb the trapped compounds into
the GC column by heating the trap to between 170 and 180°C while backflushing with
carrier gas at 20 to 60 mL/min for 4 minutes. Start MS data acquisition upon start of the
desorb cycle, and start the GC column temperature program 3 minutes later. Table 3
summarizes the recommended operating conditions for the gas chromatograph. Included
in this table are retention times and minimum levels that can be achieved under these
conditions. An example of the separations achieved by the column listed is shown in
Figure 9. Other columns may be used provided the requirements in Section 8 are met.
27

-------
If the priority pollutant gases produce GC peaks so broad that the precision and recovery
specifications (Section 8.2) cannot be met, the column may be cooled to ambient or
subambient temperatures to sharpen these peaks.
10.9	After desorbing the sample for four minutes, recondition the trap by purging with purge
gas while maintaining the trap temperature at between 170 and 180°C. After
approximately 7 minutes, turn off the trap heater to stop the gas flow through the trap.
When cool, the trap is ready for the next sample.
10.10	While analysis of the desorbed compounds proceeds, remove and clean the purge device.
Rinse with tap water, clean with detergent and water, rinse with tap and distilled water,
and dry for aminimum of 1 hour in an oven at a temperature greater than 150°C.
System performance
At the beginning of each 8 hour shift during which analyses are performed, system
calibration and performance shall be verified for the pollutants and labeled compounds
(Table 1). For these tests, analysis of the aqueous performance standard (Section 6.7.2)
shall be used to verify all performance criteria. Adjustment and/or recalibration (per
Section 7) shall be performed until all performance criteria are met. Only after all
performance criteria are met may blanks and samples be analyzed.
BFB spectrum validity: The criteria in Table 4 shall be met.
Retention times: The absolute retention times of the internal standards shall be as follows:
bromochloromethane: 653 to 782 seconds; 2-bromo-l-chloropropane: 1270 to 1369 seconds;
I,4-dichlorobutane:	1510 to 1605 seconds. The relative retention times of all pollutants
and labeled compounds shall fall within the limits given in Table 3.
GC resolution: The valley height between toluene and toluene-d8 (at m/z and 99
plotted on the same graph) shall be less than 10% of the taller of the two peaks.
Calibration verification and ongoing precision and accuracy: Compute the concentration
of each pollutant (Table 1) by isotope dilution (Section 7.4) for those compounds which
have labeled analogs. Compute the concentration of each pollutant which has no labeled
analog by the internal standard method (Section 7.5). Compute the concentrations of the
labeled compounds themselves by the internal standard method. These concentrations
are computed based on the calibration data determined in Section 7.
II.5.1	For each pollutant and labeled compound, compare the concentration with the
corresponding limit for ongoing accuracy in Table 6. If all compounds meet the
acceptance criteria, system performance is acceptable and analysis of blanks and
samples may continue. If any individual value falls outside the range given,
system performance is unacceptable for that compound.
NOTE: The large number of compounds in Table 6 present a substantial probability
that one or more will fail the acceptance criteria when all compounds are
analyzed. To determine if the analytical system is out of control, or if the
failure may be attributed to probability, proceed as follows:
11.5.1.1 Analyze a second aliquot of the aqueous performance standard (Section
6.7.2).
28
11.
11.1
11.2
11.3
11.4
11.5

-------
11.5.1.2 Compute the concentration for only those compounds which failed the
first test (Section 11.5.1). If these compounds now pass, system
performance is acceptable for all compounds, and analyses of blanks
and samples may proceed. If, however, any of the compounds fail
again, the measurement system is not performing properly for these
compounds. In this event, locate and correct the problem or recalibrate
the system (Section 7), and repeat the entire test (Section 11.1) for all
compounds.
11.5.2 Add results which pass the specification in Section 11.5.1.2 to initial (Section 8.2)
and previous on-going data. Update QC charts to form a graphic representation
of laboratory performance (Figure 8). Develop a statement of accuracy for each
pollutant and labeled compound by calculating the average percent recovery (R)
and the standard deviation of percent recovery (sr). Express the accuracy as a
recovery interval from R - 2sr to R + 2sr. For example, if R = 95% and sr = 5%,
the accuracy is 85 to 105%.
12. Qualitative determination
Identification is accomplished by comparison of data from analysis of a sample or blank
with data stored in the mass-spectral libraries. For compounds for which the relative
retention times and mass spectra are known, identification is confirmed per Sections 12.1
and 12.2. For unidentified GC peaks, the spectrum is compared to spectra in the
EPA/NIH mass spectral file per Section 12.3.
12.1	Labeled compounds and pollutants having no labeled analog (Tables 1 and 2):
12.1.1	The signals for all characteristic m/z's stored in the spectral library (Section 7.2.3)
shall be present and shall maximize within the same two consecutive scans.
12.1.2	Either (1) the background corrected EICP areas or (2) the corrected relative
intensities of the mass spectral peaks at the GC peak maximum shall agree within
a factor of 2 (0.5 to 2 times) for all masses stored in the library.
12.1.3	In order for the compounds for which the system has been calibrated (Table 1) to
be identified, their relative retention times shall be within the retention-time
windows specified in Table 3.
12.1.4	The system has not been calibrated for the compounds listed in Table 2; however,
the relative retention times and mass spectra of these compounds are known.
Therefore, for a compound in Table 2 to be identified, its relative retention time
must fall within a retention-time window of ±60 seconds or ±20 scans (whichever
is greater) of the nominal retention time of the compound specified in Table 3.
12.2	Pollutants having a labeled analog (Table 1):
12.2.1	The signals for all characteristic m/z's stored in the spectral library (Section 7.2.3)
shall be present and shall maximize within the same two consecutive scans.
12.2.2	Either (1) the background corrected EICP areas or (2) the corrected relative
intensities of the mass spectral peaks at the GC peak maximum shall agree within
a factor of two for all masses stored in the spectral library.
29

-------
12.2.3 The relative retention time between the pollutant and its labeled analog shall be
within the windows specified in Table 3.
12.3	Unidentified GC peaks.
12.3.1	The signals for m/z's specific to a GC peak shall all maximize within the same
two consecutive scans.
12.3.2	Either (1) the background corrected EICP areas or (2) the corrected relative
intensities of the mass spectral peaks at the GC peak maximum shall agree within
a factor of 2 with the masses stored in the EPA/NIH mass-spectral file.
12.4	The m/z's present in the sample mass spectrum that are not present in the reference mass
spectrum shall be accounted for by contaminant or background ions. If the sample mass
spectrum is contaminated, or if identification is ambiguous, an experienced spectrometrist
(Section 1.4) is to determine the presence or absence of the compound.
13. Quantitative determination
13.1	Isotope dilution: Because the pollutant and its labeled analog exhibit the same effects
upon purging, desorption, and gas chromatography, correction for recovery of the
pollutant can be made by adding a known amount of a labeled compound to every
sample prior to purging. Relative response (RR) values for sample mixtures are used in
conjunction with the calibration curves described in Section 7.4 to determine
concentrations directly, so long as labeled compound spiking levels are constant. For the
toluene example given in Figure 7 (Section 7.4.3), RR would be equal to 1.174. For this
RR value, the toluene calibration curve given in Figure 6 indicates a concentration of 31.8
\ig/L.
13.2	Internal standard: For the compounds for which the system was calibrated (Table 1)
according to Section 7.5, use the response factor determined during the calibration to
calculate the concentration from the following equation.
Concentration = ^
(Aix x RF)
where the terms are as defined in Section 7.5.1. For the compounds for which the system
was not calibrated (Table 2), use the response factors in Table 5 to calculate the
concentration.
13.3 The concentration of the pollutant in the solid phase of the sample is computed using the
concentration of the pollutant detected in the aqueous solution, as follows:
„ ^ ^	Ti,n\ 0.005 L x aqueous cone (iig/L)
Concentration in solid \]ig/kg) = 	3	—-
0.01 x percent solids(g)
where
"percent solids" is from Section 10.1.3
13.4 Dilution of samples: If the EICP area at the quantitation m/z exceeds the
calibration range of the system, samples are diluted by successive factors of 10
until the area is within the calibration range.
30

-------
13.4.1	For aqueous samples, bring 0.50 mL, 0.050 mL, 0.0050 mL, etc., to 5-mL volume
with reagent water and analyze per Section 10.4.
13.4.2	For samples containing high solids, substitute 0.50 or 0.050 g in Section 10.5.2 to
achieve a factor of 10 or 100 dilution, respectively.
13.4.3	If dilution of high solids samples by greater than a factor of 100 is required, then
extract the sample with methanol, as described in Section 10.6.
13.5	Dilution of samples containing high concentrations of compounds not in Table 1: When
the EICP area of the quantitation m/z of a compound to be identified per Section 12.3
exceeds the linear range of the GCMS system, or when any peak in the mass spectrum
is saturated, dilute the sample per Sections 13.4.1 through 13.4.3.
13.6	Report results for all pollutants, labeled compounds, and tentatively identified compounds
found in all standards, blanks, and samples to three significant figures. For samples
containing less than 1% solids, the units are Hg/L; and for undiluted samples containing
1% solids or greater, units are |ig/kg.
13.6.1 Results for samples which have been diluted are reported at the least dilute level
at which the area at the quantitation m/z is within the calibration range (Section
13.4), or at which no m/z in the spectrum is saturated (Section 13.5). For
compounds having a labeled analog, results are reported at the least dilute level
at which the area at the quantitation m/z is within the calibration range (Section
13.4) and the labeled compound recovery is within the normal range for the
method (Section 14.2).
14. Analysis of complex samples
14.1	Some samples may contain high levels (>1000 |ig/kg) of the compounds of interest and
of interfering compounds. Some samples will foam excessively when purged. Others will
overload the trap or the GC column.
14.2	When the recovery of any labeled compound is outside the range given in Table 6, dilute
0.5 mL of samples containing less than 1% solids, or 0.5 g of samples containing 1% solids
or greater, with 4.5 mL of reagent water and analyze this diluted sample. If the recovery
remains outside of the range for this diluted sample, the aqueous performance standard
shall be analyzed (Section 11) and calibration verified (Section 11.5). If the recovery for
the labeled compound in the aqueous performance standard is outside the range given
in Table 6, the analytical system is out of control. In this case, the instrument shall be
repaired, the performance specifications in Section 11 shall be met, and the analysis of the
undiluted sample shall be repeated. If the recovery for the aqueous performance standard
is within the range given in Table 6, then the method does not apply to the sample being
analyzed, and the result may not be reported for regulatory compliance purposes.
14.3	When a high level of the pollutant is present, reverse search computer programs may
misinterpret the spectrum of chromatographically unresolved pollutant and labeled
compound pairs with overlapping spectra. Examine each chromatogram for peaks greater
than the height of the internal standard peaks. These peaks can obscure the compounds
of interest.
31

-------
15. Method performance
15.1	The specifications for this method were taken from the interlaboratory validation of EPA
Method 624 (Reference 10). Method 1624 has been shown to yield slightly better
performance on treated effluents than method 624. Results of initial tests of this method
at a purge temperature of 80°C can be found in Reference 11 and results of initial tests
of this method on municipal sludge can be found in Reference 12.
15.2	A chromatogram of the 20 Hg/L aqueous performance standards (Sections 6.7.2 and 11.1)
is shown in Figure 9.
Reference
1.	"Performance Tests for the Evaluation of Computerized Gas Chromatography/Mass
Spectrometry Equipment and Laboratories," USEPA, EMSL Cincinnati, OH 45268,
EPA-600/4-80-025 (April 1980).
2.	Bellar, T. A. and Lichtenberg, J. J., "Journal American Water Works Association," 66, 739
(1974).
3.	Bellar, T. A. and Lichtenberg, J. J., "Semi-Automated Headspace Analysis of Drinking
Waters and Industrial Waters for Purgeable Volatile Organic Compounds," in Measurement
of Organic Pollutants in Water and Wastewater, C. E. VanHall, ed., American Society for
Testing Materials, Philadelphia, PA, Special Technical Publication 686, (1978).
4.	National Standard Reference Data System, "Mass Spectral Tape Format," U.S. National
Bureau of Standards (1979 and later attachments).
5.	"Working with Carcinogens," DHEW, PHS, NIOSH, Publication 77-206 (1977).
6.	"OSHA Safety and Health Standards, General Industry," 29 CFR 1910, OSHA 2206, (1976).
7.	"Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
Committee on Chemical Safety (1979).
8.	"Handbook of Analytical Quality Control in Water and Wastewater Laboratories," USEPA,
EMSL Cincinnati, OH 45268, EPA-4-79-019 (March 1979).
9.	"Methods 330.4 and 330.5 for Total Residual Chlorine," USEPA, EMSL Cincinnati, OH
45268, EPA-4-79-020 (March 1979).
10.	"Method 624-Purgeables", 40 CFR Part 136 (49 FR 43234), 26 October 1984.
11.	"Narrative for SAS 106: Development of an Isotope Dilution GC/MS Method for Hot
Purge and-Trap Volatiles Analysis," S-CUBED Division of Maxwell Laboratories, Inc.,
Prepared for W. A. Telliard, Industrial Technology Division (WH-552), USEPA, 401 M St.
SW, Washington DC 20460 (July 1986).
12.	Colby, Bruce N. and Ryan, Philip W., "Initial Evaluation of Methods 1634 and 1635 for the
Analysis of Municipal Wastewater Treatment Sludges by Isotope Dilution GCMS," Pacific
Analytical Inc., Prepared for W. A. Telliard, Industrial Technology Division (WH-552),
USEPA, 401 M St. SW, Washington DC 20460 (July 1986).
32

-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists
532 allyl alcohol
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
42
30
43
39
44
232
45
12
53
13
55
59
56
58
57
1000
58
300
61
15




533 carbon disulfide
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
44
282
46
10
64
14
76
1000
77
27
78
82
534 2-chloro-l,3-butadiene (chloroprene)
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
48
21
49
91
50
223
51
246
52
241
53
1000
54
41
61
30
62
54
63
11
64
16
73
21
87
12
88
452
89
22
90
137




535 chloroacetonitrile
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
47
135
48
1000
49
88
50
294
51
12
73
22
74
43
75
884
76
39
77
278




536 3-chloropropene
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
35
39
36
40
40
44
42
206
47
40
58
35
49
176
51
64
52
31
61
29
73
22
75
138
76
1000
77
74
78
324






537 crotonaldehyde
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
35
26
40
28
42
339
43
48
44
335
49
27
50
40
51
20
52
21
53
31
55
55
68
24
69
511
70
1000
71
43






33

-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists (continued)
538 1,2-dibromoethane (EDB)
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
79
50
80
13
31
51
82
15
93
54
95
42
105
32
106
29
107
1000
108
38
109
922
110
19
186
13
188
27
190
13






539 dibromomethane
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
43
99
44
101
45
30
79
184
80
35
81
175
91
142
92
61
93
1000
94
64
95
875
160
18
172
375
173
14
174
719
175
12
176
342


540 trans-l,4-dichloro-2-butene
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
49
166
50
171
51
289
52
85
53
878
54
273
62
286
64
91
75
1000
77
323
88
246
89
415
90
93
91
129
124
138
126
86
128
12


541 1,3-dichloropropane
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
40
15
42
44
47
19
48
20
49
193
51
55
61
18
62
22
63
131
65
38
75
47
76
1000
77
46
78
310
79
12






542 cis-l,3-dichloropropene
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
37
262
38
269
39
998
49
596
51
189
75
1000
77
328
110
254
112
161






543 ethyl cyanide
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
44
115
50
34
51
166
52
190
53
127
54
1000
55
193










544 ethyl methacrylate
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
m/z
Int.
42
127
43
48
45
155
55
32
58
39
68
60
69
1000
70
83
71
25
85
14
86
169
87
21
96
17
99
93
113
11
114
119




34

-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists (continued)

545 2-hexanone (methvl butvl ketone)


m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
42
61
43
1000
44
24
55
12
57
130
58
382
59
21
71
36
85
37
100
56




546 iodomethane
m/z
int.
m/z
int.
m/z
int.
m/z.
int.
m/z
int.
m/z
int.
44
57
127
328
128
17
139
39
140
34
141
120
142
1000
143
12








547 isobutyl alcohol
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
34
21
35
13
36
13
37
11
39
10
42
575
43
1000
44
42
45
21
55
40
56
37
57
21
59
25
73
12
74
63






548 methacrylonitrile
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
38
24
39
21
41
26
42
100
49
19
50
60
51
214
52
446
53
19
62
24
63
59
64
136
65
55
66
400
67
1000
68
51




549 methi
/I methacrylate
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
42
127
43
52
45
48
53
30
55
100
56
49
59
124
68
28
69
1000
70
51
82
26
85
45
98
20
99
89
100
442
101
22




550 4-methyl-2-pentanone (methyl isoboutyl ketone; MIBK)
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
42
69
43
1000
44
54
53
11
55
15
56
13
57
205
58
346
59
20
67
12
69
10
85
96
100
94










551 1,1,1,2-tetrachloroethane
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
47
144
49
163
60
303
61
330
62
98
82
45
84
31
95
416
96
152
97
270
98
84
117
804
121
236
131
1000
133
955
135
301




35

-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists (continued)

552 trichlorofluoromethane


m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
44
95
47
153
49
43
51
21
52
14
66
162
68
53
82
40
84
28
101
1000
102
10
103
671
105
102
117
16
119
14






553 1,2,3-trichloropro
pane
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
49
285
51
87
61
300
62
107
63
98
75
1000
76
38
77
302
83
23
96
29
97
166
98
20
99
103
110
265
111
28
112
164
114
25


554 vinyl acetate
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
36
5
42
103
43
1000
44
70
45
8
86
57
951 m-xylene
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
65
62
77
124
91
1000
105
245
106
580


951 0- + p-xylene
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
m/z
int.
51
88
77
131
91
1000
105
229
106
515


36

-------
OPTIONAL
FOAM TRAP
EXIT 1/4 IN. O.D,
10 MM GLASS FRIT
MEDIUM POROSITY
EXIT 1/4 IN. O.D.
14 MM O.D.
NLET 1/4 IN, O.D.


SAMPLE INLET
? WAY SYRINGE VALVE
17 CM 20 GAUGE SYRINGE NEEDLE
6 MM O.D. RUBBER SEPTUM
INLET 1/4 IN. O.D.
1/16 IN, O.D.
y STAINLESS STEEL
13X
MOLECULAR SIEVE
PURGE GAS FILTER
PURGE GAS
FLOW CONTROL
FIGURE 1
Purging Device for Waters

-------
PURGE INLET FITTING
SAMPLE OUTLET FITTING
3" x 6 MM O.O. GLASS TUBING

Hi
40 ML VIAL

k;
SEPTUM
CAP
FIGURE 2 Purging Device for Soils or Waters
38

-------
PACKING DETAIL	CONSTRUCTION DETAIL
5 MM GLASS WOOL
7.7 CM SILICA GEL
15 CM TENAX GC
•- 1 CM OV-1
5 MM GLASS WOOL
TRAP INLET
COMPRESSION
FITTING NUT
AND FERRULES
14 FT. 70/FOOT
RESISTANCE WIRE
WRAPPED SOLID
THERMOCOUPLE/
CONTROLLER
SENSOR
ELECTRONIC
TEMPERATURE
CONTROL AND
PYROMETER
TUBING 25 CM
0.105 IN. I.D.
0.125 IN. O.D.
STAINLESS STEEL
FIGURE 3 Trap Construction and Packings
39

-------
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE GAS
FLOW CONTROL
1'iX MOLECULAR
SIEVE FILTER
fj PORT
VALVE
VENT
LIQUID INJECTION PORTS
i— COLUMN OVEN
B-JUUU*-,
§"\JULrLr^r
CONFIRMATORY COLUMN
TO DETECTOR
ANALYTICAL COLUMN
OPTIONAL 4 PORT COLUMN
SELECTION VALVE
TRAP INLET
TRAP
22 X
PURGING
DEVICE
NOTE„
ALL LINES BETWEEN TRAP
AND GC SHOULD BE HEATED
TO 80"C
FIGURE 4 Schematic of Purge and Trap
Device-Purge Mode
40

-------
CARRIER GAS
FLOW CONTROL
PRESSURE
regulator
liquid injection ports
	COLUMN OVEN
CONFIRMATORY COLUMN
TO DETECTOR
PURGE GAS
FLOW CONTROL
13X MOLECULAR V/
SIEVE FILTER
6-PORT
VALVE
VENT
PURGING
DEVICE
ANALYTICAL COLUMN
OPTIONAL 4-PORT COLUMN
SELECTION VALVE
TRAP INLET
TRAP
200 "C
NOTE:
ALL LINES BETWEEN TRAP
AND GC SHOULD BE HEATED
TO 80°C.
FIGURE 5 Schematic of Purge and Trap
Device—Desorb Mods
41

-------
HI
tfi
z
o
CL
tfi
Lit
cc
LU
>
Lit
oc
i	1	r
10 20 50
CONCENTRATION (ug/L)
FIGURE 6 Relative Response Calibration Curve for
Toluene. The Dotted Lines Enclose a +/- 10 Percent
Error Window
42

-------
AREA 168920
(A)
M/Z 100
M/Z 92
(B)
AREA=60960

M/Z 100
M/Z 92
M/Z 92 = 96868
M/Z 100 82508
(C)
M/Z 100
M/Z 92
FIGURE 7 Extracted Ion Current Profiles for (A)
Toluene, (B) Toluene-ds, and (C) a Mixture of
Toluene and Toluene-ds
43

-------
N
<
<
ID
DC
<
*
<
LLl
OL
120,000
100,000
80,000
TOLUENE-D
+ 3s
ANALYSIS NUMBER
LU Q
% lii
a. 3
W J
Hi o
eg
LU uJ
> z
I- LU
53
TOLUENE
0.90
t i i i	1	1	1	1	r
6/1 6/1 6/1 6/1 6/2 6/2 6/3 6/3 6/4 6/5
DATE ANALYZED
FIGURE 8 Quality Control Charts Showing Area
(top graph) and Relative Response of Toluene to
ToIuene-d8 (lower graph) Plotted as Function of
Time or Analysis Number
44

-------
MASS CHROMATOGRAM	DATA: U0A101945 II
09/01/84 23:65:00	CALI: M0AID1945 II
SAMPLE: UO,S,OPR,00020,00,U,NA:NA,NAI
CONDS.: 16248,3.011,2HM, 3645,45-24088, i58240,20HL/MINS
RANGE: G 1,1200 LABEL: N 0, 4.0 QUANi A 0, 1.0 J
SCANS
1 TO 1208
0 BASE: U 20, 3
100.0-1
47
251
u
222376.
46.514
250.575
400
13:40
600
20:30
800
27:20
1000
34:10
=t
1200 SCAN
41:00 TIME
FIGURE 9 Chromatogram of Aqueous Performance Standard
45

-------