Example Procedure for Field-Filtering Samples for
Cryptosporidium Analysis Using
IDEXX Filta-Max™ Filters
June 2003
Draft

-------
Example Procedure for Field-Filtering Samples for Cryptosporidium
Analysis Using IDEXX Filta-Max™ Filters
1.0 Required Materials
Check to make sure the following materials are available before collecting your sample:
~	Several pairs of new, powder-free latex gloves (Lab Safety Supply, cat. number 16285XL, or
equivalent)
~	Sample collection form
~	Filta-Max™ foam filter module (IDEXX, cat. number, FMC 10603) with housing (IDEXX, cat.
number, FMC 10504)
~	Electric peristaltic pump (Masterflex L/S variable-speed drive, Cole Parmer, cat. number LC-
07554-90, or equivalent with Masterflex L/S Easy Load II pump head, Cole Parmer, cat. number
LC-77200-60, or equivalent) (for use only for collection of samples from an unpressurized
source)
~	One length of peristaltic pump tubing (Masterflex C-Flex LS 18 tubing, Cole Parmer, cat. number
EW-06424-18, or equivalent)
~	Five lengths of 12.7-mm (0.5-in.) internal-diameter clear laboratory tubing (Tygon formula R-
3603 [Cole Parmer, cat. number, 06408-18], or equivalent)
~	Five pairs of clamps fit to tubing
~	One coupling to fit 0.5-in. internal diameter tubing
~	Pressure regulator (Watts Regulator, cat. number, 263A, or equivalent)
~	Pressure gauge (maximum operating pressure 120 psi) (Cole Parmer, cat. number, E-68004-05, or
equivalent)
~	Water meter (flow totalizer) (Omega Engineering, cat. number, FTB-4005, or equivalent)
~	Flow rate meter (Blue-White, cat. number, F44375LEA 8, or equivalent)
~	Flow control valve (Plast-o-matic, model number #FC050B-l/2-PV)
~	Sample number label
~	Cooler, approximately 16-quart (Coleman, model number, 6216-703, or equivalent)
~	Two large plastic trash bags
~	One 8-lb bag of ice
~	Gel ice pack (VWR, cat. number, 15715-105, or equivalent)
~	Three 13x18 inch ziplock bag (Associated Bag, cat. number, 270-42, or equivalent)
~	Strapping tape
~	Two, self-adhesive plastic airbill sleeves
~	Airbill for shipment
Draft
1
June 2003

-------
2.0 Collecting the Sample
If the sample will be collected from a pressurized source, use the sample collection procedures in Section
2.1. If the sample will be collected from an unpressurized source, use the sample collection procedures in
Section 2.2.
2.1 Sample Filtration from a Pressurized Source
2.1.1	Before connecting the sampling system to the tap or source, turn on the tap and allow the water to
flow for 2 to 3 minutes or until any debris that has accumulated in the source water lines has
cleared or the turbidity in the water becomes visibly uniform. Turn off the tap.
2.1.2	Put on a pair of powder-free latex gloves to prevent contamination from outside sources. Any
contamination of the sampling apparatus may bias the final results.
2.1.3	Determine the pressure of the water supply using a pressure gauge.
2.1.4	Assemble the sampling system, minus the Filta-Max™ filter. For high pressure (> 20 psig) sites,
the sampling system should be assembled at the sample port valve in the following order, as
shown in Figure 1 below:
~	Reinforced influent tubing
~	Pressure regulator
~	Pressure gauge
~	Reinforced inlet tubing
~	Filta-Max™ filter housing
~	Reinforce outlet tubing
~	Flow totalizer (mechanical or graduated collection device)
~	Flow control valve or flow rate meter with valve
~	Effluent tubing to drain
Figure 1. Sample System Setup for Collecting Cryptosporidium Samples from a Pressurized
Source ( > 20 psig)
Flow rate meter*-
with valve
Pressure
gauge
Outlet tubing
Effluent tubing
\
influent Pressure Inlet tubing
tubing regulator
Filta-Max™
filter housing
Flow totalizer
\\ Flow control
\ J valve may be
	~
DIRECTION OF FLOW
used in place of
flow rale meter
Draft
2
June 2003

-------
For a low-pressure (7.5 to 20 psig) site, the sampling system should be assembled at the sample
port valve in the following order, as shown in Figure 2 below:
~	Influent tubing
~	Filta-Max™ filter housing
~	Outlet tubing
~	Flow rate meter with valve
~	Flow totalizer (mechanical or graduated collection device)
~	Effluent tubing to drain
Figure 2. Sample System Setup for Collecting Cryptosporidium Samples from a Pressurized
Source (7.5 - 20 psig)

	•• ¦4a> jcfc
/
Flow rate meter
with valve
/
Flow totalizer	Effluent tubing

Su
tubing	Filta-Max™ filter housing tubing
	(Valve)
Li	W
Influent	I	Outlet
DIRECTION OF FLOW
2.1.5	Connect the sampling system, with an empty Filta-Max™ filter housing, to the pressurized water
system. Verify that the filter housing is installed so that the end closest to the screw top cap is the
inlet and the opposite end is the outlet.
2.1.6	Slowly turn the tap to fully open. Allow a minimum of 20 L to flush the system. During this
period, perform the following steps:
2.1.6.1	Measuring flow rate with the flow rate meter or the flow totalizer and a stopwatch, adjust
the flow rate to approximately 3-4 L/min (approximately 0.8 - 1 gpm). At high pressure
sites, using the pressure regulator, adjust the pressure to a maximum of 120 psi. Observe
system for leaks and take the necessary corrective action if any are present. A differential
pressure of 7.5 psi is required to create flow through the filter. The recommended
differential pressure to produce the flow rate of 3 to 4 L/min is 75 psi. Do not exceed the
maximum operating pressure of 120 psig.
2.1.6.2	Record the following information on the sample collection form:
% Public water system (PWS) name
% PWS address
Sampler name
% Sample ID (optional)
Draft
3
June 2003

-------
%	Public Water System Identification (PWS ID) number
%	Public Water System facility ID number
%	Facility name
%	Sample collection point ID
%	Sample collection point name
%	Sample collection date
%	Source water type (optional [but required for E. coli sample forms])
% Requested analysis (circle Cryptosporidium field sample for routine monitoring
sample; circle both "Cryptosporidium field sample" and "Cryptosporidium
matrix spike" sample if you are sending an additional sample with the monitoring
sample for matrix spike analysis)
2.1.6.3 After the system has been flushed, any optional water quality parameters such as
temperature, pH, and/or turbidity should be measured now.
2.1.7	Turn off the water at the sample port valve when the flow rate has been adjusted and the system
has been flushed.
2.1.8	Record the following information on the sample collection form:
% Current meter reading
2.1.9	Install the Filta-Max™ filter into the housing and secure the housing cap by hand tightening.
Apply gentle pressure to create a seal between the module and the "O" rings in the base and lid of
the housing. Excessive tightening is not necessary, and may shorten the life of the "O" rings. A
light application of vacuum grease may be used to lubricate the "O" rings, but too much grease
will produce a negative effect.
Note! Retain the end caps provided with the filter housing. These caps will be needed to seal
the housing for shipment.
2.1.10	Turn on the pressurized water source. Adjust flow to within 3 to 4 L/min, if necessary.
% Record start time on the sample collection form.
2.1.11	Monitor the water meter. When the targeted volume (typically 10 L [2.64 gal] or 5 0 L [ 13.2 gal])
has passed through the Filta-Max™ filter, shut off the water source. Allow the pressure to
decrease until the water stops.
2.1.12	Record the following information on the sample collection form:
% Stop time (when the water was shut off)
% Final meter reading
% Comments to laboratory, if needed
2.1.13	Disconnect the inlet end of the filter housing, making sure not to spill any of the water remaining
in the capsule. This water is part of your sample.
Draft
4
June 2003

-------
2.1.14	The filter can either be shipped in the filter housing or removed from the filter housing and
shipped alone.
2.1.15	If the filter will be shipped without the housing, open the housing and dump the filter and any
water remaining in the housing into a ziplock bag. Place this bag inside a second ziplock bag and
seal.
2.1.16	If the filter will be shipped in the housing, seal the inlet and outlet of the housing with the rubber
stoppers that were previously saved. Place the filter housing containing the filter in a plastic
ziplock bag for shipment.
2.1.17	Place a label on the outer ziplock bag containing the filter and using a waterproof pen record the
following information:
^ PWS ID
% Facility name
% Date of sample collection
Cover the label with clear tape to keep the label dry.
2.1.18	Immediately following sample collection, place the bag containing the filter (with or without
filter housing) in a refrigerator to chill prior to packing the shipping cooler for shipment. If no
refrigerator is available, and the sample will not be shipped for several hours, place the bag in the
shipping cooler with ice to chill, and replace the ice before shipping.
Note! Method 1622/1623 requires that the temperature of the sample upon arrival at the
laboratory must be <10°C (but not frozen), and the laboratory must have the filter
processed within 96 hours of sample collection. If the sample temperature and holding
time requirements are not met, then the sample is invalid and must be recollected.
As a recommendation, it is best to have the filter processed by the laboratory within 24 hours of
sampling.
2.2 Sample Filtration from an Unpressurized Source
2.2.1	Put on a pair of powder-free latex gloves to prevent contamination from outside sources. Any
contamination of the sampling apparatus may bias the final results.
2.2.2	If sampling from a source of unlimited volume, it may be desirable to pre-flush the sampling
system. Assemble the sampling system, minus the Filta-Max™ filter. Verify that the filter
housing is installed so that the end closest to the screw top cap is the inlet and the opposite end is
the outlet. The sampling system should be assembled in the following order, as shown in Figure
3 below:
~	Influent tubing
~	Peristaltic pump
~	Inlet tubing
~	Filta-Max™ filter housing
~	Outlet tubing
Draft
5
June 2003

-------
~	Flow rate meter with valve
~	Flow totalizer (mechanical or graduated container)
~	Effluent tubing to drain
Figure 3. Sample System Setup for Collecting Cryptosporidium Samples from an Unpressurized
Source
Flow rate
meter
with valve
3	
/
FIow totalizer Effluent tubing
Peristaltic pump
Filta-Max™ filter housing
m
|—(Valve)
Influent tubing
Inlet tubing
Outlet tubing
Sample
	~
DIRECTION OF FLOW
2.2.3	Place the inlet end of the inlet tubing in sample source, away from any walls, bottom, or other
environmental surfaces.
2.2.4	Turn on the pump and allow a minimum of 20 L to flush the system. If sampling source water
from a carboy, continuously refill the carboy as necessary to flush the system. As a
recommendation, the carboy should not be removed from the sampling chain and should be
refilled using a separate container. Observe the system for leaks and take the necessary corrective
action if any are present. During this period, perform the following steps:
2.2.4.1	Measuring flow rate with the flow rate meter or the flow totalizer and a stopwatch, adjust
the flow rate to approximately 3-4 L/min (approximately 0.8-1 gpm) by varying the
pump speed or adjusting the valve (if pump is not variable speed).
2.2.4.2	Record the following information on the sample collection form:
% Public water system (PWS) name
%	PWS address
Sampler name
%	Sample ID (optional)
%	Public Water System Identification (PWS ID) number
%	Public Water System facility ID number
%	Facility name
%	Sample collection point ID
%	Sample collection point name
%	Sample collection date
Draft
6
June 2003

-------
% Source water type (optional [but required for E. coli sample forms])
% Requested analysis (circle Cryptosporidium field sample for routine monitoring
sample; circle both "Cryptosporidium field sample" and "Cryptosporidium
matrix spike" sample if you are sending an additional sample with the monitoring
sample for matrix spike analysis)
2.2.4.3 Once the system has been flushed, any optional water quality parameters such as
temperature, pH, and/or turbidity should be measured now.
2.2.5	Turn off the pump when the flow rate has been adjusted and the system has been flushed.
Following flushing, or if flushing is not performed, proceed with the following steps:
2.2.6	Record the following information on the sample collection form:
% Current meter reading
2.2.7	Install the Filta-Max™ filter into the housing and secure the housing cap by hand tightening.
Apply gentle pressure to create a seal between the module and the "O" rings in the base and lid of
the housing. Excessive tightening is not necessary, and may shorten the life of the "O" rings. A
light application of vacuum grease may be used to lubricate the "O" rings, but too much grease
will produce a negative effect.
Note! Retain the end caps provided with the filter housing. These caps will be needed to seal
the housing for shipment.
2.2.8	Turn on the peristaltic pump. Record the following on the sample collection form:
% Start time
2.2.9	Monitor the water meter. When the targeted volume (typically 10 L [2.64 gal] or 50 L [13.2 gal])
has passed through the Filta-Max™ filter or the container is depleted, remove the inlet tubing
from the source. Shut off the pump when all the water has been forced through the filter. Allow
the pressure to decrease until the water stops. If sampling from a container, add 1 L of reagent
water per every 10-L container volume and rinse the sides of the container. Resume pumping
until the rinse volume is filtered.
2.2.10	Record the following information on the sample collection form:
% Stop time (when the pump was shut off)
% Final meter reading or measured total volume sample, not including reagent rinse
volume, if any
% Comments to laboratory, if needed
2.2.11	Disconnect the inlet end of the filter housing, making sure not to spill any of the water remaining
in the capsule. This water is part of your sample.
2.2.12	The filter can either be shipped in the filter housing or removed from the filter housing and
shipped alone.
Draft
1
June 2003

-------
2.2.13	If the filter will be shipped without the housing, open the housing and dump the filter and any
water remaining in the housing into a ziplock bag. Place this bag inside a second ziplock bag and
seal.
2.2.14	If the filter will be shipped in the housing, seal the inlet and outlet of the housing with the rubber
stoppers that were saved previously. Place the filter housing containing the filter in a plastic
ziplock bag for shipment.
2.2.15	Place a label on the outer ziplock bag containing the filter and using a waterproof pen record the
following information:
^ PWS ID
% Facility name
% Date of sample collection
Cover the label with clear tape to keep the label dry.
2.2.16	Immediately following sample collection, place the bag containing the filter housing in a
refrigerator to chill prior to packing the shipping cooler for shipment. If no refrigerator is
available, and the sample will not be shipped for several hours, place the bag in the shipping
cooler with ice to chill. Replace the ice before shipping.
Note! Method 1622/1623 requires that the temperature of the sample upon arrival at the
laboratory must be <10°C (but not frozen), and the laboratory must have the filter
processed within 96 hours of sample collection. If the sample temperature and holding
time requirements are not met, then the sample is invalid and must be recollected.
As a recommendation, it is best to have the filter processed by the laboratory within 24 hours of
sampling.
3.0	Packing the Sample
3.1	Insert two large plastic trash bags into the shipping cooler to create a double liner. Immediately
before packing the cooler, create two 4-lb ice packs by putting half of an 8-lb bag of ice into two
separate ziplock bags. To prevent leaks place each ice pack into an additional ziplock bag. Gel
packs or blue ice may be used instead of wet ice, as long as the sample is maintained in the
appropriate temperature range. Seal the ziplock bag, expelling as much air as possible, and secure
top with tape.
Note! Shipping companies may delay sample shipments if leakage occurs. Double liners and
ziplock bags around ice will prevent leakage and delays.
3.2	Place the bag containing the filter housing into the shipping container. Place the ice pack around,
but not on, the sample bag to prevent freezing. Inflated, empty sample bags can be placed
between the filter housing and the ice packs to prevent the filter from freezing.
3.3	If you will be monitoring sample temperature during shipment, place in the cooler the
temperature monitoring device (e.g., extra sample bottle for measuring sample temperature upon
receipt at the laboratory, thermometer vial, or Thermochron™ iButton). Seal each liner bag by
twisting top of bag and tying in a knot.
Draft
8
June 2003

-------
3.4	Peel the backing off one of the plastic airbill sleeves and attach the sleeve to the inside of the
cooler lid.
% Sign and date the sample collection form
Fold the completed sample collection form, and place it inside the plastic sleeve.
3.5	Close the cooler lid, seal the horizontal joints with duct tape, and secure the lid with tape by
taping the cooler at each end, perpendicular to the seal.
Note! Shipping companies may delay sample shipments if leakage occurs. Be sure to seal the
cooler joints.
3.6	Peel the backing off of the second airbill sleeve and attach the sleeve to the outside of the cooler
lid. Complete the shipping airbill with the laboratory address, billing information, sample weight,
and shipping service. Remove the shipper's copy of the airbill, and place the remainder of the
airbill inside the plastic sleeve.
4.0	Shipping and Tracking
4.1	Ship samples on the day of collection and use a reliable shipping service for next-day delivery. If
samples are not shipped the day of collection, the sample must be maintained at <10°C (but not
frozen) by chilling in a refrigerator or cooler filled with ice.
4.2	Contact the laboratory to notify them of the sample shipment. Request that the laboratory contact
you the next day if the sample is not received.
4.3	Using the airbill number on the shipper's copy of the airbill, track the sample shipment using the
shipping company's web page or by contacting the shipping company over the phone.
4.4	If problems are encountered with the shipment, communicate with the shipping company to
resolve, and update the laboratory regarding the status of the shipment.
Draft
9
June 2003

-------