EPA 600/R-15/076 I May 2015 I www.epa.gov/research
United States
Environmental Protection
Agency
Effect of Pressure Washing
Conditions on the Removal of Cs
from Urban Surfaces
ASSESSMENT AND EVALUATION REPORT
Office of Research and Development
National Homeland Security Research Center

-------
Effect of Pressure Washing Conditions on the
Removal of Cs from Urban Surfaces
Assessment and Evaluation Report
National Homeland Security Research Center
Office of Research and Development
U.S. Environmental Protection Agency
Research Triangle Park, NC 27711
ii

-------
Disclaimer
The U.S. Environmental Protection Agency, through its Office of Research and Development, funded and
managed the research described here under Contract #EP-C-09-027 to ARCADIS US, Inc. It has been
subjected to the Agency's review and has been approved for publication. Note that approval does not
signify that the contents necessarily reflect the views of the Agency. Mention of trade names, products, or
services does not convey official EPA approval, endorsement, or recommendation.
Questions concerning this document or its application should be addressed to:
Sang Don Lee, Ph.D.
Decontamination and Consequence Management Division
National Homeland Security Research Center
U.S. Environmental Protection Agency (E-311K)
Office of Research and Development
109 T.W. Alexander Drive
Research Triangle Park, NC 27711
Phone: 919-541-4531
Fax: 919-541-0496
E-mail: lee.sanadon@epa.gov

-------
Acknowledgments
Contributions of the following individuals and organizations to this report are gratefully acknowledged:
Peer Reviewers
Matthew Magnuson, U.S. Environmental Protection Agency, Office of Research and Development
Tom Mahler, U.S. Environmental Protection Agency, Region 7
Charles Hooper, U.S. Environmental Protection Agency, Region 7
ARC AD IS US, Inc.
iv

-------
Table of Contents
Disclaimer	iii
Acknowledgments	iv
List of Figures	vi
List of Tables	vii
List of Acronyms and Abbreviations	vii
Executive Summary	ix
1	Introduction	1
2	Materials and Methods	2
2.1	Test Overview	2
2.2	Building Materials	2
2.3	Coupon Conditioning	3
2.4	Cesium Particle Deposition	3
2.5	Pressure Washing System	5
2.6	Wash Down Conditions	7
2.6.1	Pressure Washing	7
2.6.2	Water Pressure	8
2.6.3	Wash Pattern	8
2.7	Test Matrix	8
2.8	Analysis of Pressure Wash Rinsates	9
2.8.1	Rinsate Sample Preparation	10
2.8.2	Solid Sample Preparation	10
2.8.3	Sample Analysis	11
2.8.4	Laboratory Data Entry, Validation, and Reporting	12
2.8.5	Laboratory Calculations	12
2.9	Cs Removal Efficacy	12
2.10	Statistical Methods	13
2.10.1	P-value	13
2.10.2	ANCOVA Analysis	13
3	Quality Assurance/Quality Control	14
3.1	Calibration	14
3.1.1	Pressure Plate Calibration	14
3.1.2	ICP-MS Calibration	15
3.2	Moisture Measurements	15
3.3	Quality Control Samples	16
4	Results	18
4.1 Cesium Removal Efficacy	18
v

-------
4.1.1	Overview: Brick	18
4.1.2	Overview: Concrete	21
4.1.3	Overview: Asphalt	24
4.1.4	Effects of Water Pressure	26
5	Summary and Recommendations	31
6	References	32
List of Figures
Figure 2-1. Finished concrete (a), brick (b), and asphalt (c) coupons	2
Figure 2-2. Particle deposition onto a brick and concrete coupon	4
Figure 2-3. Particle deposition onto an asphalt coupon	4
Figure 2-4. High-pressure washer system	5
Figure 2-5. Pressure washing chamber	6
Figure 2-6. Asphalt coupon in the chamber holder	6
Figure 2-7. Horizontal coupon being sprayed in chamber	7
Figure 2-8. Water wash down test pattern	8
Figure 4-1. Removal efficacy from brick - horizontal orientation and wet deposition	19
Figure 4-2. Removal efficacy from brick - vertical orientation and wet deposition	19
Figure 4-3. Removal efficacy from brick - horizontal orientation and dry deposition	20
Figure 4-4. Removal efficacy from brick - vertical orientation and dry deposition	20
Figure 4-5. Removal efficacy from concrete - horizontal orientation and wet deposition	22
Figure 4-6. Removal efficacy from concrete - vertical orientation and wet deposition	22
Figure 4-7. Removal efficacy from concrete - horizontal orientation and dry deposition	23
Figure 4-8. Removal efficacy from concrete - vertical orientation and dry deposition	23
Figure 4-9. Removal efficacy from asphalt - horizontal orientation and wet deposition	25
Figure 4-10. Removal efficacy from asphalt - horizontal orientation and dry deposition	25
Figure 4-11. Total removal efficacy comparison for horizontal orientation and dry deposition	26
Figure 4-12. Total removal efficacy comparison for horizontal orientation and wet deposition	26
Figure 4-13. Solid efficacy comparison for horizontal orientation and dry deposition	27
Figure 4-14. Solid efficacy comparison for horizontal orientation and wet deposition	28
Figure 4-15. Liquid efficacy comparison for horizontal orientation and dry deposition	29
Figure 4-16. Liquid efficacy comparison for horizontal orientation and wet deposition	29
Figure 4-17. Total efficacy comparison for vertical orientation and wet/dry deposition	30
vi

-------
List of Tables
Table 2-1.	Building Material Descriptions and Sources	2
Table 2-2.	Test Matrix for Pressure Washing Tests	9
Table 2-3.	Typical Operating Conditions for ELAN DRC-e	10
Table 3-1.	Horizontal Pressure Plate Calibration Tests	14
Table 3-2.	Vertical Pressure Plate Calibration Tests	14
Table 4-1.	Test Conditions and Removal Efficacy Results for Brick	18
Table 4-2.	Test Conditions and Removal Efficacy Results for Concrete	21
Table 4-3.	Test Conditions and Removal Efficacy Results for Asphalt	24
List of Acronyms and Abbreviations
ASTM	American Society for Testing and Materials, now
ASTM International
ANCOVA	Analysis of Covariance (Model)
ANOVA	Analysis of Variance (Model)
Ba	Barium
Ce	Cerium
CeO	Cerium Oxide
Co	Cobalt
COTS	Commercial Off-The-Shelf
Cs	Cesium
CsCI	Cesium Chloride
Dl	Deionized
EPA	U.S. Environmental Protection Agency
gpm	Gallon(s) per Minute
GLM	General Linear Model
HNO3	Nitric Acid
HSRP	Homeland Security Research Program
ICP-MS	Inductively Coupled Plasma - Mass Spectrometry
(meter)
ID	Identification
ITSD	Internal Standard
LIMS	Laboratory Information Management System
MB	Method Blank
Mg	Magnesium
NHSRC	National Homeland Security Research Center
Pb	lead
PDS	Post Digestion Spike
psi	Pound(s) per square inch
vii

-------
QA
Quality Assurance
QAPP
Quality Assurance Project Plan
QC
Quality Control
QCS
Quality Control Standard
RDD
Radiological Dispersal Device
RH
Relative Humidity
Rh
Rhodium
RSD
Relative Standard Deviation
Sr
Strontium

-------
Executive Summary
The U.S. Environmental Protection Agency's (EPA) Homeland Security Research Program (HSRP) helps
protect human health and the environment from adverse impacts of terrorist acts by carrying out
performance tests on homeland security technologies. This study investigated the impact of water
pressure conditions for decontamination of urban surfaces contaminated with cesium (Cs). The
contaminated surfaces were prepared with two different deposition methods to control the penetration
depth of Cs using water and methanol as solvents; the surfaces were then washed in a chamber,
simulating the delivery of high pressure water. Various water pressures, applied in both vertical and
horizontal orientations of the surface, were evaluated for efficacy against the contaminated coupon
surfaces. The amount of Cs in the liquid and solid portions of the water rinsate samples was measured
and analyzed to determine the impact of individual wash conditions.
The main goal of this project was to determine the impact of water pressure conditions in the
decontamination of urban surfaces contaminated with Cs particles. Various conditions of water pressure,
deposition type, and surface orientation were tested to investigate the impact on Cs removal from three
different urban surfaces. Increased water pressure improved Cs removal efficacy for all three surfaces.
Total removal efficacy varied by surface type and deposition method. Asphalt showed the highest
removal efficacy for Cs (42 % to 78 %), followed by brick (5 % to 35 %), and concrete (6 % to 25 %).
Increased Cs removal was observed from dry (methanol) deposition as opposed to wet (water) deposition
for all three surfaces. This increased Cs removal may be related to the penetration depth of Cs after
deposition. Methanol evaporates more quickly than water, and as a result, the majority of the Cs remains
close to the surface and becomes easier to remove. Orientation was not a significant factor for either
liquid or solid removal for all three surfaces. The Cs removal mechanism was further investigated by
separate measurement of the amount removed in solid phase and liquid phase.
The results showed that when comparing Cs amount in solid samples for all three substrates for the same
conditions, asphalt showed decreasing Cs amount with increasing pressure, but brick and concrete
showed increasing Cs amount with increasing pressure. Comparison of the Cs amount in liquid samples
for the same conditions showed the same profound effect for asphalt as total efficacy, whereas brick and
concrete showed a minimal increase in Cs amount with increasing pressure.
The dry and wet deposition test results demonstrated that delayed pressure washing may lead to
decreased removal of Cs from the surface due to the subsurface penetration. For asphalt, Cs removal in
the liquid phase was dominant, and the removal efficacy increased with increased water pressure up to
7000 psi. The water pressure (6000 to 7000 psi) provided high decontamination efficacy (50 % to 80 %)
with minimal surface degradation. In the case of brick and concrete, test results showed a minimal
increase (less than 10 % increase) in Cs removal as a function of pressure in the range of 4000 to
7000 psi. However, analysis of the data showed that increased removal efficacy for brick and concrete
were related to removal of solid materials by the high pressure stream. Extensive layer removal from brick
and concrete surfaces is expected to increase the removal efficacy because of subsurface penetration
and Cs sorption on the removed surfaces.
ix

-------
1 Introduction
A radiological dispersal device (RDD), also called a dirty bomb, is the combination of a conventional
explosive device with radioactive materials that can be obtained from industrial, commercial, medical, or
research applications.12 An RDD attack can impact a society in various ways including creation of
casualties, disruption of the economy, and relocation of people from the contaminated area.3 4 5 Fast and
cost-effective decontamination strategies are critical to minimize the social and economic damage from
an RDD event, increasing the community resilience.
The U.S. Environmental Protection Agency (EPA) has conducted a series of tests to investigate the fate
and transport of RDD materials on urban surfaces under varied conditions.6 7 8 9 10 The results from these
previous studies showed varied fates for cesium (Cs) on surfaces, depending on contaminant deposition
conditions, surface types, and environmental conditions. A follow-up study11 focused on the assessment
of various conditions of water wash down parameters, including wash down duration, water pressure,
wash angle, and wash patterns on asphalt, brick, and concrete using cesium chloride (133CsCI) particles.
The results showed a positive correlation with wash duration and water pressure for the asphalt and
concrete samples and demonstrated that a 90-degree wash angle was more effective than a 45-degree
wash angle for the brick and concrete samples. However, these findings are applicable primarily to the
conditions where Cs particles have limited penetration into porous surfaces.
The current study tested water pressure in an effort to optimize the pressure washing methods to
increase the removal efficacy on these same three urban surface materials (asphalt, brick, and concrete).
The research focused on the assessment of various water pressures for Cs removal as functions of
deposition methods, surface types, and water application orientations.
1

-------
2 Materials and Methods
2.1	Test Overview
The amounts of Cs removed from urban surfaces via high pressure washing were studied under various
application conditions. Various water pressures, Cs deposition methods, and vertical and horizontal
surface orientations were explored. Three different building materials were contaminated with Cs particles
via aerosolization, and the contaminated surfaces were washed using a high pressure washer. The liquid
and solid (filtered) portions of the rinsate samples were analyzed for Cs amounts that were removed from
the surface.
2.2	Building Materials
Building materials are described in Table 2-1. Concrete coupons (12" x 12" x 1.5" [Wx L x H]) were
prepared according to the method used in the previous Water Wash Down study.11 Each coupon was
visually inspected and any coupons with defects, cracks, or stains were discarded. Concrete coupons
were prepared following the manufacture instructions (Quikrete® Portland Cement Atlanta, GA). Brick
coupons (12" x 12" x 2.25" [Wx L x H]) were prepared by a brick mason. The brick walls were built using
mortar and with the same method employed when constructing a home. Asphalt coupons (6" diameter)
were supplied by the North Carolina Department of Transportation. Since the asphalt coupons were from
actual roadways, thickness varied, but test coupons were chosen to be as close to a height of 1.5 inches
as possible. Figure 2-1 shows the finished concrete, brick, and asphalt coupons.
Table 2-1. Building Material Descriptions and Sources
Material
Description
Locality
Source
Brick
Red, fine-grained
Made from North Carolina red
Triassic clay
Triangle Brick Company,
Durham, North Carolina
Concrete
Sand, Portland cement mix
N/A
Home Depot, North Carolina
Asphalt
Laboratory pressed asphalt
N/A
North Carolina Department of
Transportation
Figure 2-1. Finished concrete (a), brick (b), and asphalt (c) coupons
Five sides of each concrete and brick coupon and the edges and back of the asphalt coupons were
sealed with water-impermeable sealant (Stonelok™ E3, Richard James Specialty Chemicals Corp.,
Hastings on Hudson, NY). The top surface remained unsealed for deposition of surrogate RDD
contaminants and application of pressurized water. The sample identification (ID) was marked on two
opposite sides and the back of each coupon with paint markers before sealing.
2

-------
2.3	Coupon Conditioning
Concrete coupons were poured and cured under ambient conditions for at least 28 days before being
prepared for surface contamination as described below. Prior to particle deposition, the top surface of all
coupons was cleaned thoroughly with a 2300 psi/2.3 gallons per minute (gpm) pressure washer to
remove any loose pieces of building material. After cleaning, the coupons were dried in laboratory air for
one week. Daily moisture measurements were then taken for one week after the coupons were allowed to
dry in the laboratory (Dawson Multi-Function Moisture Meter, Model DSM170, Diamond Bar, CA). Minimal
differences in moisture were observed and appeared to be a function of room relative humidity (RH). The
moisture measurements are discussed in Section 3.2. After deposition, coupons were stored in laboratory
air (23±2 °C and 40±2 % RH) for 24 hours prior to use in the pressure washer testing.
2.4	Cesium Particle Deposition
A solution containing Cs particles in deionized (Dl) water (wet) or methanol (dry) was deposited onto
coupons using a metered syringe (MicroSprayer® Aerosolizer, Model 1A-1C and FMJ-250 High Pressure
Syringe, Penn Century, PA). Dry Cs deposition mimics the surface contamination status immediately after
contamination and wet Cs deposition represents the surface contamination condition after rain or high
relative humidity. It is expected that dry deposition causes less Cs subsurface penetration than wet
deposition due to fast evaporation rate of methanol compared to water. Templates were used to ensure
that the Cs was deposited over the majority of the surface of the coupon in a fixed-array pattern. The tip
of the high pressure syringe was positioned 5 inches from all coupon surfaces during deposition. To
determine the optimum deposition height, multiple tests were conducted using food coloring in water and
methanol. The colored solution was deposited on white paper at various heights. It was desired to have
the deposition pattern be well dispersed, but not producing a spray pattern that impacted the sides of the
deposition template, nor producing a semi-direct deposition pattern onto the white paper. Figure 2-2
shows the nine-hole template used for the 12-inch by 12-inch concrete and brick coupons. The openings
are pinhole sized, but the shape of the grid can be inferred from the location of the high pressure syringe
tips placed on the diagonal. Figure 2-3 shows the two-hole template used for the six-inch diameter
asphalt coupons. In all cases, each template hole was dosed with 25 |jl_ of the 200 ppm CsCI solution
(wet or dry) and allowed to dry under ambient conditions for 24 hours before testing. This procedure
yields a nominal deposition of 45 jjg of CsCI onto the concrete and brick coupons and 10 jjg of CsCI onto
the asphalt coupons. The deposition amount was calibrated (spike control) by depositing the cesium
solutions onto a 5 x 5 array of clean polyethylene plastic sheets with the syringe tip at the same distance
as the building material coupons. Each set of spike control samples for Cs was transferred to a clean
1000-mL polyethylene beaker for extraction. The beaker was filled with 1 % ultrapure OPTIMA nitric acid
(HNO3) (Sigma-Aldrich®, St. Louis, MO) in Dl water until the solution covered the plastic surface entirely
(~360 mL). The spike control samples were extracted by sonication for 20 minutes. After sonication, the
plastic squares were removed and discarded. The beaker was filled to 450 mL with HNO3. Using an
Eppendorf pipette, 0.5 mL was extracted from the beaker and placed into a clean 50-mL vial. The vial
was then filled to a volume of 50 mL with Dl water and submitted for analysis by Inductively Coupled
Plasma - Mass Spectrometry (ICP-MS).
3

-------
Figure 2-2. Particle deposition onto a brick and concrete coupon
Figure 2-3. Particle deposition onto an asphalt coupon

-------
2.5 Pressure Washing System
The contaminated coupons were subjected to pressure washing using a high-pressure washer system
(Silver Eagle Manufacturing, PO Box 7346, Cut-N-Shoot, Texas, 77306; Engine Driven High Pressure
Washer System) within 24 hours after contamination. Triplicate coupons were prepared for each
pressure-washing test condition for brick and concrete. Six coupons were prepared for each pressure-
washing test condition for asphalt. After a test, coupons were allowed to dry in laboratory air.
All coupons were washed in a chamber using a high-pressure water-delivery system (Figure 2-4) and
25-degree fan angle for each nozzle. A stainless steel chamber (4' x 4' x 4') was used and is shown in
Figure 2-5. The nozzle was fixed in the center of a chamber side and chamber top to accommodate both
vertical and horizontal coupon holders that were positioned on the bottom or back wall of the chamber. In
Figure 2-5, a brick coupon can be seen in the horizontal position; the coupon holder for the vertical
orientation coupons can be seen on the back wall. Figure 2-6 shows an asphalt coupon in its holder. The
front door of the chamber was opened to allow cleaning of the inside of the chamber after each test. The
chamber bottom was slanted on all four sides to collect the rinsate water. To ensure accurate and
consistent delivery of pressure during testing, the chamber was fitted with two pressure plates, one each
in the horizontal and the vertical orientation. At the beginning and end of each day of testing, the pressure
measurement for each nozzle and water pressure being tested that day.
Figure 2-4. High-pressure washer system
5

-------
Figure 2-5. Pressure washing chamber
Figure 2-6. Asphalt coupon in the chamber holder

-------
Dl water was used for the pressure-washer tests, and the water was applied for 10 seconds at a distance
often inches from the coupon surface. The rinsate water was collected in clean plastic containers
(Cubitainer®, Fisher Scientific, Pittsburgh, PA) located under the bottom drain. The rinsate containers
were weighed to record the amount of water collected during each pressure-washer test; the containers
were then capped and submitted to the laboratory for filtration and analysis of both the solid and liquid
portions by ICP-MS. After the test, the coupons were removed from the chamber coupon holder, and the
chamber surfaces were pressure-washed with a ten-second Dl water rinse. This blank chamber rinse was
also collected, and a 50-rriL vial was submitted to the laboratory for Cs analysis. As an additional
precaution against cross-contamination, the inside chamber was further rinsed with Dl water using a
garden hose for 25 to 30 seconds, taking care to clean out any particulate matter remaining around the
drain. This final rinse was collected in a five-gallon bucket and discarded.
2.6 Wash Down Conditions
2.6.1 Pressure Washing
All coupons were washed for a duration of 10±1 seconds and at a 90-degree angle between water jet
direction and coupon surface. Coupons were tested in both vertical and horizontal orientations for brick
and concrete; asphalt was tested only in the horizontal orientation. Figure 2-7 shows a brick coupon being
sprayed from the top of the chamber. Other test conditions were deposition method (wet or dry) and the
water pressure (psi) being tested. Water from the pressure washer covered the entire coupon surface.
The rinsate volume for the ten-second washes was collected into a single one-gallon Cubitainer. The
rinsate samples were analyzed for amount of Cs to determine whether the decontamination efficacy is
affected by the coupon orientation at various water pressures, using both wet and dry deposition methods
and coupon orientation.
Figure 2-7. Horizontal coupon being sprayed in chamber
7

-------
2.6.2 Water Pressure
Five different water pressures were applied to the coupons from the same distance and duration. The
applied nozzle water pressures were 2100, 4000, 5000, 6000, and 7000 psi. Pressure was adjusted by
changing the nozzle type. The 2100 psi/2.2 gpm pressure was achieved by using a Commercial Off-The-
Shelf (COTS) pressure washer purchased from Lowes (Troy-Bilt, Briggs and Stratton Power Products,
PO Box 702, Milwaukee, Wl). The high pressure washer used high pressure ceramic nozzles, and the
flow ranged from 4.5 to 7.9 gpm. The rinsates from varied water pressure tests were analyzed for Cs
amount to determine the dependence of decontamination efficacy on water pressure.
2.6.3 Wash Pattern
A single wash pattern was tested on all coupons. Figure 2-8 shows the pattern proceeding from top to
bottom. The pressure wash duration was ten seconds, wash angle was 90 degrees, and distance from
the nozzle tip to the coupon surface was ten inches.
Figure 2-8. Water wash down test pattern
2.7 Test Matrix
The coupon test matrix is shown in Table 2-2. Three samples of each building material were prepared for
both deposition methods and coupon orientation, with the exception of asphalt. Since asphalt is
encountered in real-life use only in the horizontal orientation, the vertical orientation was not tested for
this material. As shown in Table 2-2, a total of three blank coupons were prepared for each test scenario,
except six coupons were prepared for asphalt. Blank coupons were coupons without Cs deposition.
These blank coupons were washed down and the rinsate samples were collected for baseline Cs solid
and liquid concentration determination.
8

-------
Table 2-2. Test Matrix for Pressure Washing Tests
Material
Deposition
Condition
Orientation
Building Material Samples

Wet
Horizontal
3

Dry
Horizontal
3
Fine Aggregate Concrete
Wet
Vertical
3
Dry
Vertical
3

Blank
Horizontal
3

Blank
Vertical
3

Wet
Horizontal
3

Dry
Horizontal
3
Brick Wall
Wet
Vertical
3
Dry
Vertical
3

Blank
Horizontal
3

Blank
Vertical
3

Wet
Horizontal
6
Asphalt
Dry
Horizontal
6

Blank
Horizontal
6
2.8 Analysis of Pressure Wash Rinsates
Standard methods modified by the laboratory for this work included EPA Method 6020A, "Inductively
Coupled Plasma-Mass Spectrometry" and EPA Method 3050B "Acid Digestion of Sediments, Sludges
and Soils". The custom methodology consisting of the analysis of rinse water samples via ICP-MS was
developed by Pace Analytical Services, Inc. (6701 Conference Drive, Raleigh, NC 27607) and
implemented by First Analytical Laboratories (7517-101 Precision Drive, Raleigh, NC 27617) to determine
and quantify the presence of ultra-trace elemental content of Cs, cobalt (Co), and strontium (Sr). A
custom methodology consisting of the analysis of particulate matter obtained from the filtration of rinse
water samples via ICP-MS was developed and implemented to determine and quantify the presence of
ultra-trace elemental content, specifically Cs. Results were reported for use in the determination of rinse
procedure efficacy when carried out utilizing various rinsing techniques on an assortment of substrate
types.
The test rinsate samples, in one-gallon high-density polyethylene Cubitainers, were delivered in their
entirety to the analytical laboratory (First Analytical Labs, Raleigh, NC). For the blank chamber rinsate
samples, a 50-mL polyethylene centrifuge tube was submitted, and the remainder of the blank rinsate
sample was discarded. The samples were subjected to a visual validity assessment by the laboratory to
ensure that no samples were compromised during transit (i.e., leaked sample, loose sample container lid,
cracked sample container, etc.). Following the validity check, the samples were compared to the total
number of samples denoted on the chain-of-custody form to ensure that all samples were present at time
of receipt.
9

-------
First Analytical Labs analyzed both the rinsate and filtered (solid) portions of each sample using a custom
methodology developed by Pace Analytical Services (Raleigh, NC) for the ultra-trace analysis of Cs by
ICP-MS: "ICP-MS Analysis of Filtered Rinse Water for Ultra-Trace Levels of Cesium (Cs), Cobalt (Co)
and Strontium (Sr), RAL-0001-MET, Revision 1" and "ICP-MS Analysis of Particulate Matter Obtained
From Filtered Rinse Water for Ultra-Trace Levels of Cesium (Cs), RAL-0002-MET, Revision-New". These
methods were modified versions of EPA Standard Methods 200.8 and 6020A and are summarized in
more detail below. The modified methods were verified by the laboratory before use. A model ELAN
DRC-e ICP-MS (Perkin Elmer, Concord, Ontario) was used for Cs analysis. The typical operating
conditions of ICP-MS are summarized in Table 2-3.
Table 2-3. Typical Operating Conditions for ELAN DRC-e
Parameter
Operating Condition
Pulse Stage Voltage
1700 Volts
Analog Stage Voltage
-2100 Volts
Radio Frequency Generator Power
1200 Watts
Lens Voltage
7 Volts
Vacuum Pressure
4.5 e -005 Torr
Nebulizer Gas Flow Rate
0.96 L/min
Auxiliary Gas Flow Rate
15.0 L/min
Nebulizer Type
Ryton Scott Cross Flow Spray Chamber with
Alumina Injector Tube
Ambient Temperature
72.0 °F ± 5.0 °F (22.2 °C ± 2.8 °C)
Sampler and Skimmer Cone Type
Platinum
Mass Spectrometer Type
Quadrupole
Detector Type
Dual Mode Electron Multiplier
2.8.1	Rinsate Sample Preparation
Following the custody sign-over process, the samples were logged into the First Analytical laboratory
information management system (LIMS). Each rinsate sample went through seven main process phases
once received and logged into the LIMS: (1) sample preparation, (2) sample analysis, (3) data
interpretation, (4) data entry, (5) preliminary data submission, (6) data validation, and (7) final report
generation/distribution.
2.8.2	Solid Sample Preparation
Each solid sample went through nine main process phases once received and logged into the LIMS:
(1) sample filtration, (2) sample digestion, (3) sample digest preparation, (4) sample analysis, (5) data
interpretation, (6) data entry, (7) preliminary data submission, (8) data validation, and (9) final report
generation/distribution.
Prior to beginning the analytical process, the samples were filtered using a standard manual vacuum
system to separate the solid portion of the sample from the aqueous portion of the sample. The liquid
10

-------
obtained from the filtration process was retained and stored in high-density polyethylene sample
containers for subsequent analysis as described in Section 2.8.3. The solid material obtained from the
filtration process was collected on acid-washed filter papers and placed in a laboratory oven at 100±10 °C
to dry for a minimum of four hours.
Upon completion of the drying process, the filters with the retained solids were removed from the oven
and allowed to cool to ambient temperature under a clean fume hood. Dried and cooled filters were
ground into a consistency that was more easily digested and weighed out into pre-labeled digestion vials
using an analytical balance. In addition to the samples, there were two digestion vials labeled "Method
Blank (MB)" and "Post Digestion Spike (PDS)". The vial labeled MB was prepared by weighing out 10
grams of ASTM International (ASTM) Type I water, and the vial labeled PDS was prepared by cutting up
one raw filter and using an analytical balance to determine the weight of the raw filter added to the vial. All
sample weights were recorded on each digestion vial as well as on the applicable sample preparation
bench sheet.
All digestion vials were then placed into a polycarbonate digestion rack, covered with a watch lid, and
transported to a fume hood. While under the fume hood, 10 mL of a 1:1 ultra-trace metal grade nitric
acid/ASTM Type I water solution was added using an adjustable 1-to-10-mL pipette, re-covered with
watch lids, and placed into a block digestion system (Environmental Express, Charleston, SC, Model
SC-100) pre-heated to 105±1 °C. The samples were heated for approximately 30 minutes, removed from
the block digestion system, and allowed to cool to room temperature under a closed sash fume hood.
Once cooled, 5 mL of ~70-percent ultra-trace metal grade nitric acid (HNO3) was added to each vial, and
the rack of covered samples was returned to the block digestion system and pre-heated to a temperature
of 95±1 °C. The samples were refluxed until all brown fumes subsided, adding additional 5-mL aliquots of
HNO3 as necessary. Once all brown fumes were absent, the samples were removed from the block
digestion system and allowed to cool to room temperature, after which 2 mL of ASTM Type I water and
2 mL of hydrogen peroxide (H2O2) were added to each vial, respectively, and the covered vials were
returned to the block digestion system and allowed to heat for 30 minutes at a temperature of 95±1 °C.
After a period of 30 minutes, the samples were visually checked for signs of effervescence. Should such
signs be observed, additional water and H2O2 were added to the vials in 2-mL aliquots, and the samples
were allowed to heat for additional 30 minute intervals. This process was continued until no signs of
effervescence remained.
Upon completion of the effervescent phase, the samples were heated on a hot plate for a period of 60
minutes, making sure that samples did not reduce to dryness. The samples were cooled to ambient
conditions under a closed sash fume hood and then filtered using clean funnels lined with quantitative-
grade, ashless filter paper and washed with ASTM Type I water. The filtrate was collected into a clean
labeled vial and brought to a final volume of 50 mL using ASTM Type I water. Prior to capping the
samples, the PDS vial was spiked with 50 |jL of the secondary source Cs certified reference material. The
samples were then transferred for analysis via ICP-MS as described in Section 2.8.3.
2.8.3 Sample A nalysis
Following the completion of a successful calibration and quality control (QC) check standard routine (see
Section 3.1), each batch of samples and controls was loaded onto the ICP-MS for analysis. The samples
were prepared by pipetting 10 mL of a well-mixed sample, as well as 10 |jL of an internal standard
solution, and then dispensing both liquids into a pre-labeled polyethylene test tube. After all samples
11

-------
within a given batch were prepared, all test tubes were capped and shaken thoroughly to ensure even
dispersion of the internal standard solution within the sample aliquot. In addition to the preparation of
batched samples, the appropriate number of control and duplicate samples were prepared and included in
the batch prior to loading the samples onto the instrument for analysis.
Any sample that yielded a result greater than the highest standard on the calibration curve was re-
analyzed using the lowest dilution factor necessary to bring the result to a concentration within the
calibration curve range. Any sample that yielded a result significantly impacted by matrix interferences
was repeated using the lowest dilution factor necessary to reduce or eliminate the interfering
component(s) to a level where the potential for erroneous signal amplification or suppression was
mitigated. All internal standard recovery values were monitored to ensure that there were no sample
introduction issues, instrument drift, or other problems that have the potential to effect data quality.
2.8.4	Laboratory Data Entry, Validation, and Reporting
All sample and QC data were entered into a customized Excel-based spreadsheet. All sample- and
batch-related QC data were included.
2.8.5	Laboratory Calculations
All internal standard (USD) recovery calculations used the following formula:
[USD (Sample)/ITSD (Calibration Blank)] * 100
All quality control standard (QCS) isotopic recovery calculations used the following formula:
[QCSobtained Conc./QCSlarget Cone. ]* 100
All sample/duplicate relative percent difference calculations used the following formula:
[(Resulti - Result2)/ResultAvg] * 100
All solid sample results were calculated using the following formula:
Result (|jg/Kg) = [Result (mg/L) * 50/ Sample Aliquot (g)] * 1000
2.9 Cs Removal Efficacy
The efficacy of a pressure washer test was assessed by determining the amount of Cs in the water wash
rinsate samples. The Cs amount in the rinsate water samples was compared to the Cs amount in the
positive control rinsate. Removal efficacy of Cs from the coupon material was calculated as the ratio of Mr
and Mpc:
Total Cs Removal Efficacy (%) = (Msr+ Mlr) / Mpc x 100
where Msr and Mlr are the average Cs amounts (jjg) in solid and liquid phases in triplicate rinsate
samples, respectively, and Mpc is the average Cs amount (jjg) from five positive controls.
12

-------
2.10 Statistical Methods
Statistical analyses were performed on the data set to determine statistical significance of results. The
specific tests performed are discussed below.
2.10.1	P-value
The p-value is defined as the probability of obtaining a result equal to or more extreme than the value
actually observed. The p-values calculated for these analyses were set at the 95 % confidence level.
Therefore, if an obtained p-value is lower than 0.05, the set of data being examined is significant at the 95
% confidence level.
2.10.2	ANCOVA Analysis
A general linear model (GLM) (Snedecor and Cochran, 1980, Statistical Methods12) was used to evaluate
the Cs removal testing. The following categorical independent variables were evaluated as factors in the
model: orientation (H, V), material (asphalt, brick, concrete), and deposition (wet, dry). These treatment
factors represent the categorical independent variables in the model. The testing was carried out at five
different pressure levels, and pressure was evaluated as a continuous independent variable. The
dependent variables evaluated were liquid Cs removed, solid Cs removed, and total Cs removed. This
modeling approach is commonly referred to as an analysis of covariance model (ANCOVA). ANCOVA
models combine the features of analysis of variance models (ANOVA) and linear regression models and
are used to evaluate treatment differences when there are both categorical and continuous treatment
factors in an experiment (Snedecor and Cochran, 1980). Both main and crossed effects were evaluated
for statistical significance. A post-hoc multiple-comparison test (Tukey [1953])13) was used to evaluate the
model-predicted means for each factor for both main and crossed effects when significant. All modeling
was conducted using the GLM procedure of SAS 9.3 statistical software (SAS Institute, Cary, NC,
201214).
13

-------
3 Quality Assurance/Quality Control
Quality assurance/quality control (QA/QC) procedures were performed according to the EPA-approved
quality assurance project plan (QAPP) for this test. All equipment and monitoring devices used at the time
of evaluation were verified as being calibrated.
3.1 Calibration
3.1.1 Pressure Plate Calibration
The wash chamber was equipped with two pressure plates, one mounted horizontally and one mounted
vertically. The pressure plates were attached to a computer program that logged the actual pressure in
pounds being exerted on the plate by the pressure washer nozzle for a given pressure. During each day
of testing, multiple tests of pressure were conducted by spraying onto the pressure plate for
approximately 20 seconds. In addition, a 25-pound kettle bell was placed on the horizontal plate to
provide a consistent measurement for each calibration test. The pressure plate calibration was performed
before and after each day of testing to document any drift in the pressure reading during the actual
testing. Each pressure plate calibration test at each pressure consisted of a minimum of 10 individual
measurements. The data collected were used to calculate an average at each pressure level for testing
and a relative standard deviation (RSD). The results of these pressure tests, for both horizontal and
vertical orientations, are presented in Tables 3-1 and 3-2, respectively.
Table 3-1. Horizontal Pressure Plate Calibration Tests

Before
After
Measurement
(lb)
Average
(lb)
RSD
(%)
Average
(lb)
RSD
(%)
25
25.02
0.4
24.98
0.4
2100
4.15
12.7
4.19
9.2
4000
20.86
16.7
22.21
8.4
5000
21.61
6.2
22.50
5.2
6000
23.17
9.2
23.47
5.5
7000
22.95
7.3
23.49
5.0
Table 3-2. Vertical Pressure Plate Calibration Tests

Before
After
Measurement
(lb)
Average
(lb)
RSD
(%)
Average
(lb)
RSD
(%)
2100
3.51
25.9
3.00
22.7
4000
22.53
4.5
21.62
9.7
5000
21.88
8.6
19.45
12.8
6000
23.10
10.7
23.26
6.1
7000
22.79
10.0
22.61
7.5
14

-------
3.1.2 ICP-MS Calibration
A daily performance assessment was carried out on the ICP-MS prior to the analysis of any samples. This
assessment was performed using a multi-element certified reference material (Perkin Elmer Smart Tune
solution, 10.0 |jg/L of various elements - Perkin Elmer Catalog #N8125040) obtained from the instrument
manufacturer. Upon completion of a valid daily performance assessment, the instrument was calibrated
for the element of interest using one of the following methods: Low Level Cesium Method, Low Level
Cobalt Method, or Low Level Strontium Method.
All of the above methods employed a four-point, external, linear-through zero calibration type with curve
points of 0.025, 0.50, 5.0 and 10.0 |jg/L; each method utilized an isotope (other than the element of
interest) to serve as an internal standard.
Prior to instrument calibration and data analysis, instrument performance was validated by performing a
daily performance assessment. The criteria for a passing daily performance assessment included:
•	Rhodium (Rh) counts per second must be > 150,000
•	Lead (Pb) counts per second must be > 100,000
•	Magnesium (Mg) counts per second must be > 20,000
•	Cerium (Ce)/Cerium Oxide (CeO) ratio should be < 3.00 %
•	Barium (Ba)++ should be < 5.00 %
•	Combined 220 mass background must be < 30 counts per second
•	RSD among all replicate sweeps for each tune component (sans 220 background) must be < 5.00 %
All calibration curve correlation coefficient numbers were > 0.9995. The calibration curve was verified at
the beginning and at the end of the analytical sequence by analyzing a blank QC standard (0.000 jjg/L)
that yielded a result < 0.025 |jg/L, a low-range QC standard (0.500 jjg/L) that yielded a result between
0.450 and 0.550 |jg/L, a mid-range QC standard (5.00 jjg/L) that yielded a result between 4.50 and 5.50
|jg/L, and a high-range QC standard (10.0 jjg/L) that yielded a result between 9.00 and 10.0 |jg/L. The
percent recovery for all applicable internal standards must be between 70 % and 125 %. Sample batch
sizes did not exceed 20 samples, and each batch of 20 samples was bracketed by one blank QC and one
low-range, mid-range, or high-range QC standard, making sure to rotate between all three range
standards with each additional batch analyzed. In addition to batch-associated QC standards, one sample
duplicate was prepared per every ten samples within the batch. The maximum relative percent difference
between the two results was < 10.00 percent.
3.2 Moisture Measurements
Moisture measurements were taken of all coupons after the initial pressure wash with the 2300 (2100) psi
washer and air drying for one week. These measurements were taken to demonstrate that the moisture
15

-------
level in the coupons was not changing overtime and to ensure that all coupons were in a "dry" state
before being tested. The measurements were taken for five consecutive days prior to any deposition and
indicated that the surfaces of the coupons were dry throughout the measurement period. The moisture of
the asphalt coupons could not be measured due to incompatibility with the moisture meter. However, the
average moisture measurements for the brick and concrete coupons were 0.13 percent and 1.49 percent,
respectively.
3.3 Quality Control Samples
QC samples generated during testing included use of positive control coupons, blank coupons, and high
pressure water blank samples. Blank coupons were coupons that had not been dosed with any
contaminant. The high pressure water blank samples were taken directly from the bottom of the chamber.
Positive controls were created by cutting Ziploc® plastic bags into three-inch squares after the bags were
washed and cleaned using Triton-X cleaning solution. These squares were placed (overlapping) on the
surface of a blank large coupon so that the entire surface of the coupon was covered with the plastic
Ziploc® material. There were a total of 25 Ziploc squares for each positive control deposition. The
deposition blank coupon was covered by a 5 x 5 array of Ziploc® squares overlapping each other. The
actual deposition area for the 3x3 array of holes in the template was 36 in2. The CsCI contaminant
solution was deposited using the high pressure syringe. After depositing the CsCI solution into all
deposition locations, all squares were placed into a clean plastic 1000-mL beaker. A plastic beaker was
used to insure no absorption of Cs on the silica in the glass beaker. A 1 % HNO3 solution was used to fill
the beaker to approximately 360 mL. Paraffin was placed on top of the beaker to prevent any loss of
HNO3; the beaker was then sonicated for 15 minutes to promote the release of Cs particles into the
solution. After sonication, each Ziploc® square was removed from the beaker and discarded. HNO3 was
added to the beaker up to a 450-mL total volume. Using an Eppendorf pipette, 0.5 mL was extracted from
the beaker and placed into a clean 50-mL vial. The vial was then filled to a volume of 50 mL with Dl water
and submitted for analysis by ICP-MS using EPA Standard Method 200.8. A modified EPA Method 200.8
was used for analysis since EPA 200.8 specifications do not list Cs as a measured component.
The average recoveries for the Cs positive controls were between 70 percent and 120 percent. The RSD
of Cs amounts from positive control recovery results was less than 19 percent. The background levels
from the blank coupons were subtracted from the test coupon results. The average Cs removed for the
brick blank coupons was 0.652 jjg with a standard deviation of 0.175. The average Cs removed for the
concrete blank coupons was 0.466 jjg with a standard deviation of 0.225. The average Cs removed for
the asphalt blank coupons was 0.061 jjg with a standard deviation of 0.047. The rinsate from the brick
blank chamber washes was analyzed for Cs, and the average Cs obtained was 0.159 jjg with a standard
deviation of 0.121. The rinsate from the concrete blank chamber washes was analyzed forCs and the
average Cs obtained was 0.155 jjg with a standard deviation of 0.150. The rinsate from the asphalt blank
chamber washes was analyzed forCs, and the average Cs obtained was 0.129 jjg with a standard
deviation of 0.128.
The ANCOVA model was also used to analyze the blank results. The results indicated orientation was not
a significant predictor for either liquid or solid removal. For liquid removal, pressure was significant for
brick only. For solid removal, pressure was significant for brick and concrete only. Mean liquid removal
was similar in brick and concrete and was greater than for asphalt. Mean solid removal was similar for
asphalt and concrete, which were both less than for brick. These results should be used with caution
16

-------
because brick and concrete blanks were not tested at the lower pressures, and ANCOVA models are
predicting negative results for solids at lower pressures.
17

-------
4 Results
4.1 Cesium Removal Efficacy
Efficacy tables and comparison plots are presented for each material in the following subsections. The
results are presented as total removal efficacy. Solid Cs amount is the amount of Cs removed from the
coupon via the solid phase of the rinsate. Liquid Cs amount is the amount of Cs removed from the
coupon via the liquid phase of the rinsate. The total Cs removal efficacy is the sum of solid Cs efficacy
(Msr) and liquid Cs efficacy (Mlr) as shown in Section 2.9.
4.1.1 Overview: Brick
Cesium removal efficacy results for brick are listed in Table 4-1. The removal efficacy results reflect the
average of three coupons. The solid and liquid Cs amount removed was blank coupon-corrected using a
bulk average blank for each material. If the resultant solid/liquid Cs amount removed resulted in a
negative value, a "zero" was used for that solid/liquid quantity.
Table 4-1. Test Conditions and Removal Efficacy Results for Brick
Coupon
ID
Nozzle Pressure
(psi)
Deposition
(Wet/Dry)
Coupon
Orientation
(HA/*)
Total Cs Removal
Efficacy
(%)
252-03
2100
Dry
H
14.2
255-03
2100
Dry
V
17.2
259-03
4000
Dry
H
30
262-03
4000
Dry
V
27.5
265-03
5000
Dry
H
34.3
268-03
5000
Dry
V
34.9
216-03
6000
Dry
H
30.1
219-03
6000
Dry
V
29.8
222-03
7000
Dry
H
31.8
225-03
7000
Dry
V
27.7
234-03
2100
Wet
V
8.3
231-03
2300
Wet
H
5.1
237-03
4000
Wet
H
14.1
240-03
4000
Wet
V
15.6
243-03
5000
Wet
H
15
246-03
5000
Wet
V
20.9
201-03
6000
Wet
H
19.3
204-03
6000
Wet
V
16.8
207-03
7000
Wet
H
20.2
210-03
7000
Wet
V
23.3
*Horizontal/Vertical
18

-------
The following figures illustrate the Cs removal efficacy results for the brick pressure washing tests.
Figures 4-1 and 4-2 show the results for the wet deposition method and the horizontal and vertical
orientations, respectively. Figures 4-3 and 4-4 show the same results for the dry deposition method. In
addition, each chart shows the error bars with one standard deviation. Note that each data point for
efficacy represents the average for a set of triplicate test results.
25

—X—.

T


>
U
(0
u
it
I 1

Liquid
5
n
HI
H
h
2100 4000 5000 600C
Pressure jpsi)
70
30
Figure 4-1. Removal efficacy from brick - horizontal orientation and wet deposition



I
I


I™ Liquid



IB Solid
1

¦
14
2100	4000	5000	6000	7000
Pressure (psi)
Figure 4-2. Removal efficacy from brick - vertical orientation and wet deposition
19

-------
40
35
30
25
20
15
10


i

I


-L














|


i

JjgSSJ


i







j





1
2100
4000
5000
Pressure (psi)
6000
Liquid
Solid
7000
Figure 4-3. Removal efficacy from brick - horizontal orientation and dry deposition
ro 20
Pressure (psi)
Figure 4-4. Removal efficacy from brick - vertical orientation and dry deposition
20

-------
4.1.2 Overview: Concrete
Cesium removal efficacy results for concrete are listed in Table 4-2. The removal efficacy results reflect
the average of three coupons. The solid and liquid Cs amount removed was blank coupon-corrected
using a bulk average blank for each material. If the resultant solid/liquid Cs amount removed resulted in a
negative value, a "zero" was used for that solid/liquid quantity.
Table 4-2. Test Conditions and Removal Efficacy Results for Concrete
Coupon ID
Nozzle
Pressure (psi)
Deposition
(Wet/Dry)
Coupon
Orientation
(H/V)
Total Cs Removal
Efficacy
(%)
122-03
2100
Dry
H
15.1
125-03
2100
Dry
V
19
129-03
4000
Dry
H
15.5
132-03
4000
Dry
V
21.5
135-03
5000
Dry
H
24.9
138-03
5000
Dry
V
10.3
160-03
6000
Dry
H
13.9
163-03
6000
Dry
V
17.8
184-03
7000
Dry
H
14.7
187-03
7000
Dry
V
24.2
101-03
2100
Wet
H
12.2
104-03
2100
Wet
V
5.6
107-03
4000
Wet
H
18
110-03
4000
Wet
V
13.4
113-03
5000
Wet
H
13.9
116-03
5000
Wet
V
16.8
144-03
6000
Wet
V
15.2
147-03
6000
Wet
H
16.6
151-03
7000
Wet
H
21.1
154-03
7000
Wet
V
14.3
The following figures illustrate the Cs removal efficacy results for the concrete pressure washing tests.
Figures 4-5 and 4-6 show the results for the wet deposition method for the horizontal and vertical
orientations, respectively. Figures 4-7 and 4-8 show the same results for the dry deposition method. In
addition, each chart shows the error bars with one standard deviation. Note that each data point for
efficacy represents the average for a set of triplicate test results.
21

-------








-¦


I

-I



I

I




Liquid




1

2100	4000	5000	6000	7000
Pressure (psi)
Figure 4-5. Removal efficacy from concrete - horizontal orientation and wet deposition
20
18
16
14
a? 12
I* 10
o
£ 8
6
4
2
0






¦¦






1	1

























1



1


















Liquid
Solid
2100
4000
5000
Pressure (psi)
6000
7000
Figure 4-6. Removal efficacy from concrete - vertical orientation and wet deposition
22

-------
2100	4000	5000	6000	7000
Pressure (psi)
Figure 4-7. Removal efficacy from concrete - horizontal orientation and dry deposition
30
25
20
NO
I" 15
u
10
2100	4000	5000	6000
Pressure (psi)
T
7000
Liquid
Solid
Figure 4-8. Removal efficacy from concrete - vertical orientation and dry deposition
23

-------
4.1.3 Overview: Asphalt
Cesium removal efficacy results for asphalt are listed in Table 4-3. The removal efficacy results reflect the
average of six coupons. The solid and liquid Cs amount removed was blank coupon-corrected using a
bulk average blank for each material. If the resultant solid/liquid Cs amount removed resulted in a
negative value, a "zero" was used for that solid/liquid quantity.
Table 4-3. Test Conditions and Removal Efficacy Results for Asphalt
Coupon ID
Nozzle
Pressure
(psi)
Deposition
(Wet/Dry)
Total Cs Removal
Efficacy
(%)
006-03
2100
Dry
45.1
034-03
4000
Dry
51.5
037-03
5000
Dry
57.3
091-03
6000
Dry
73.1
094-03
7000
Dry
78.3
001-03
2100
Wet
45.9
013-03
4000
Wet
55.1
025-03
5000
Wet
42.1
048-03
6000
Wet
55.8
086-03
7000
Wet
73.8
The following figures illustrate the Cs efficacy results for the asphalt pressure washing tests. Figures 4-9
and 4-10 show the results for the wet and dry deposition methods for the horizontal orientations,
respectively. In addition, each chart shows the error bars with one standard deviation. Note that each
data point for efficacy represents the average for a set of six test results.
24

-------
80
70
60
VP
>
50
40
30
20
10
2100
I
"
JtrpSllKU
mi
4000
i
" >:
: :':i _
I
¦ ¦
I

...
"
isf
5000
Pressure (psi)
6000
-
' :
- -
m

*=Vr-T
—
. ;
7000
Liquid
¦ Solid
Figure 4-9. Removal efficacy from asphalt - horizontal orientation and wet deposition
90
80
70
60
S?
> 50
U
(0
I 40
LU
30
20
10
0
2100
"
iwioi
4000
..
•:i» •	i

5000
Pressure (psi)
x
-
J ' • ii:
_
U" — i- <
-±-
9, -m . ¦ ¦
• ¦ . ...
"
«5 'His
-
	I	Lii	
6000
— —
: ' • : '¦! -
r.-t* ' ¦ ¦ **--r
!Lrir::
ii
IS;
ill -
w- «gp!
_
— —
i
7000
Figure 4-10. Removal efficacy from asphalt - horizontal orientation and dry deposition
25

-------
4.1.4 Effects of Water Pressure
A total efficacy comparison with a linear regression analysis (Microsoft Excel, Microsoft Corporation, One
Microsoft Way, Redmond, WA, 2013) for the Cs removal as a function of water pressure was conducted
for the three materials, and the results are shown in Figures 4-11 and 4-12 for all conditions tested. The
pressure effect via the ANCOVA model varied by material (but not by deposition method) and was
significant only for asphalt. Efficacy increase as a function of water pressure was statistically significant
for asphalt (p = 0.011) with the dry deposition method. Efficacy as a function of water pressure was not
statistically significant for brick (p = 0.106) or concrete (p = 0.969).

90

80

70

60


>
50
u

CU
40
O


LU
30

20

10

0





y = 0.0051x + 5.8625 		

R2 = 0.8236 	


	•
		


•	
y = 0.0023x +0.5589
R2 = 0.8635
y = 0.0016X + 4.9805 ¦ 		

f	*	*	i




¦ Brick
A Concrete
• Asphalt
2000
4000
Pressure (psi)
6000
8000
Figure 4-11. Total removal efficacy comparison for horizontal orientation and dry deposition
80
70
60
^ 50
5" 40
(C
lac
20
10
0



y = 0.0017x +15.432
R2 = 0.4193
•



	
!-.««•*




¦ 	
1		1
		



*	
	 1
•







y =
0.0014X-
0.2299
y =
0.0027X-2.3826
R2 = O 7466



R2 = 0.9504
B 	I


J.		
. *	
			




m	





¦ Brick
A Concrete
• Asphalt
1000 2000 3000 4000 5000
Pressure (psi)
6000
7000
8000
Figure 4-12. Total removal efficacy comparison for horizontal orientation and wet deposition
26

-------
The total removal efficacy comparison for the horizontal orientation and wet deposition method shows
that pressure was extremely significant for brick (p = 0.005), but not significant for concrete (p = 0.130)
nor asphalt (p = 0.194). Total removal efficacy (R2 = 0.75) via the ANCOVA model varied by material and
deposition and was greatest for asphalt. The pressure effect evaluated by the ANCOVA model varied by
material, but not by deposition method and was significant only for asphalt and brick.
A comparison of solid Cs efficacy for all three substrates and all conditions tested is shown in Figures
4-13 and 4-14. As shown, the solid efficacy for asphalt decreased with increasing pressure. However,
solid efficacy for brick and concrete increased with increasing pressure, most likely due to asphalt being
comprised of tar and not smaller particles like brick and concrete. Solid removal efficacy via the ANCOVA
model (R2 = 0.35) varied by material but not deposition method and was greatest in asphalt and brick.
20
18
16
14
E 12
ra 10
ru
u
it 8
LU
6
4
2
0

T \/
- 0 0015x- 4 6467


K
— W . W W _L ^ A I . w I /
R2 = 0.6766



1




\
v, _ n nrn c>, ¦ in nm
C t

y — -u.uuiuA t j
R2 = 0.89:
> * x
T
y = 0.0013x
- 3.5403

i
R2 = 0.
672
	


l	: ,


. 		
L
2000
¦ Brick
A Concrete
• Asphalt
4000
Pressure (psi)
6000
8000
Figure 4-13. Solid efficacy comparison for horizontal orientation and dry deposition
27

-------
25
20
S? 15
s-
u
(D
U
10


1_





y = I
-0.0019X + 12.891






R2 = 0.908
K
n
~

y = 0.0003X + 1.9527 m 	
n *• V . * - ¦ •*
R2 = 0.4803 " 		 1
	
«¦••• • T»*
	 % .••A
•••* 	 		 .n T



m y = 0.002x- 6.1202
R2 = 0.6986
¦ Brick
A Concrete
• Asphalt
1000 2000 3000 4000 5000 6000
Pressure (psi)
7000
8000
Figure 4-14. Solid efficacy comparison for horizontal orientation and wet deposition
A comparison of the liquid efficacy for all three substrates and all conditions tested is shown in Figures
4-15 and 4-16. The liquid efficacy increased marginally for brick, but not concrete. However, liquid
efficacy increased substantially for asphalt with increasing pressure, as opposed to the solid efficacy. For
the liquid ANCOVA model (R2 = 0.71), the main effects (material, pressure, and deposition method) were
significant. Liquid removal efficacy was greatest in asphalt and this may be due to fewer debris generated
compared to brick and concrete coupons.
28

-------

90

80

70

60
3?

s-
50
u

(0

u
40
it


30

20

10

0





y = 0.0067X - 5.0899 ,•
R2 = 0.8586 •



\




¦
\i = n nnnQv j. a >;q5r ¦••'"•

y - . -r.-,,
R2 = 0.6524

¦ V
¦¦ 0.0002X + 8.8534
11? /-V /-\A •	7

H V* ¦ r\ = U.U1D /

A —
2000
4000
Pressure (psi)
¦ Brick
A Concrete
• Asphalt
6000
8000
Figure 4-15. Liquid efficacy comparison for horizontal orientation and dry deposition
80
70
60
50
40
(T3
30
20
10
0






f |


y =
:0.0035x +2.5411
D2 _ n "71 QA














•


y = o.
I
0015x +1.3417
?2 — n 7mfi

m







y =
O.OOllx
-2.0914



X

R = 0.9172
r.-.-.-.v—>¦




¦	
		*	


¦ Brick
A Concrete
• Asphalt
0 1000 2000 3000 4000 5000 6000 7000 8000
Pressure (psi)
Figure 4-16. Liquid efficacy comparison for horizontal orientation and wet deposition
A total efficacy comparison for the vertical orientation and wet and dry deposition is shown in Figure 4-17.
For wet deposition, brick showed increasing efficacy with increasing pressure, and concrete showed
decreasing efficacy at 6000 and 7000 psi. For dry deposition, the brick efficacy decreased at 6000 and
7000 psi. The concrete efficacy for the dry deposition decreased at 5000 psi, but increased at 6000 and
7000 psi. For the vertical orientation and wet deposition method, pressure was significant for brick (p =
0.039) but not significant for concrete (p = 0.113). However, pressure was not significant for brick (p =
29

-------
0.215) nor concrete (p = 0.802) in the vertical orientation with the dry deposition method. Results from the
ANCOVA model indicated orientation was not a significant factor for either liquid or solid removal.
40
35
30
25
>
S 20
ru
u
1 c
LU 15
10
5
0
y = 0.004x + 6.4642
I

R = 0.55

I

[
y = 0.0027x +3.7986 $
r2 — n
H
M



j
	
s
y = 0.0004X + 16.4!
i		J
	\
•»•••••••••
i		
R2 = 0.0244
		
•
y = 0.
D018x + 4.2477


F
I2 = 0.6216
2000
4000
Pressure (psi)
6000
8000
¦ Brick Wet
~	Brick Dry
•	Concrete Wet
A Concrete Dry
Figure 4-17. Total efficacy comparison for vertical orientation and wet/dry deposition
30

-------
5 Summary and Recommendations
This study investigated the impact of water pressure conditions for decontamination of urban surfaces
contaminated with Cs. The study was conducted to evaluate the decontamination efficacy of pressure
washer as functions of water pressure, surface type, surface orientation, and deposition method. The
evaluation was performed by contaminating common urban surfaces with nonradioactive CsCI particles
and then washing them in a chamber simulating the delivery of high pressure water. The amount of Cs in
the liquid and solid portions of the water rinsate samples were measured and analyzed for the impact of
individual wash conditions.
The current study found the following key information for surface Cs removal using a pressure washer:
•	Increased pressure improved Cs removal efficacy for all three surfaces. Asphalt showed a much
more profound effect with increasing pressure than brick and concrete.
•	The Cs removal mechanism was further investigated by measuring removal efficacy in solid and
liquid portion separately. The results showed that when comparing solid efficacy for all three
substrates for the same conditions, asphalt showed decreasing contribution of solid efficacy with
increasing pressure, but brick and concrete showed increasing contribution of solid efficacy with
increasing pressure. The cause of negative correlation for asphalt's solid efficacy with increased
water pressure needs further investigation. Liquid efficacy comparison for the same conditions
showed the same profound effect for asphalt as total efficacy, whereas brick and concrete showed
minimal increase in efficacy with increasing pressure.
•	Total removal efficacy varied by surface types and deposition methods. Asphalt showed the highest
Cs removal efficacy (42 % to 78 %), followed by brick (5 % to 35 %) and concrete (6 % to 25 %).
Higher Cs removal was observed from dry (methanol) deposition than from wet (water) deposition for
all three surfaces, possibly related to the penetration depth of Cs after deposition. Methanol
evaporates faster than water, and as a result, the majority of Cs remains close to the surface and is
easier to remove.
•	Orientation was not a significant factor for either liquid or solid removal for all three surfaces.
The dry and wet deposition test results demonstrated that delayed application of the pressure washer
may lead to less removal of Cs from the surface due to the subsurface penetration which may increase
overtime. For asphalt, Cs removal in the liquid phase was dominant, and the removal efficacy was
increased with increased water pressure up to 7000 psi. The water pressure (6000 to 7000 psi) will
provide high decontamination efficacy (50 % to 80 %) with minimal destruction. In the case of brick and
concrete, test results showed minimal increase (less than 10 % increase) in Cs removal as a function of
pressure in the range of 4000 to 7000 psi. However, increased removal efficacy for brick and concrete
were related mainly to removal of solids from the coupon by the high pressure stream. Extensive layer
removal from brick and concrete surfaces is expected to increase the removal efficacy due to the
subsurface penetration and Cs sorption on the removed surfaces. For effective concrete and brick
surface decontamination using pressure washer, it will be necessary to assess two items: Cs subsurface
distribution and surface removal thickness as a function of water pressure.
31

-------
6 References
1	Gonzalez, A.J. (2003) Security of radioactive sources: threats and answers. In International Conference
on Security of Radioactive Sources, pp.33-58. Vienna, Austria: International Atomic Energy Agency.
2	NRC (May 2007) Backgrounder on Dirty Bombs: U.S. Nuclear Regulatory Commission.
3	Zimmerman, P.D. and Loeb, C. (2004) Dirty Bombs: The threat revisited. In Defense Horizons pp.1-11.
4	Karam, P.A. (2005) Radiological terrorism. Hum. Eco.I Risk Assess. 11, 501-523.
5	Rosoff, H. and von Winterfeldt, D. (2007) A risk and economic analysis of dirty bomb attacks on the
ports of Los Angeles and Long Beach. Risk Anal. 27, 533-546.
6	EPA 2012. Fate of Radiological Dispersal Device (RDD) Material on Urban Surfaces: Impact of Rain on
Removal of Cesium, US EPA/600/R/12/569.
7	Lee, S.D., Snyder, E.G., Willis, R., Fischer, R., Gates-Anderson, D., Sutton, M., Viani, B., Drake, J. and
MacKinney, J. (2010) Radiological dispersal device outdoor simulation test: Cesium chloride particle
characteristics. J. Hazard. Mater. 176, 56-63.
8	EPA 2010. Radiological Dispersal Device Outdoor Simulation Test: Fate of Cs on Limestone,
EPA/600/X/10/005.
9	Gusarov, A., N'icheva, N., Konoplev, A., Lee, S.D., Maslova, K., Popov, V., and Stepina, I. (2011) Fate
and transport of radiocesium in urban building materials. International Conference on Radioecology &
Environmental Radioactivity: Environment & Nuclear Renaissance 46(6): S265-S269.
10	Maslova, K., Stepina, I., Konoplev, A., Popov, V., Gusarov, A., Pankratov, F., Lee, S.D., and N'icheva,
N. (2013) Fate and transport of radiocesium, radiostrontium and radiocobalt on urban building materials,
Journal of Environmental Radioactivity 125: 74-80.
11	Water Wash Down of Radiological Dispersal Device (RDD) Material on Urban Surfaces: Effect of
Washing Conditions on Cs Removal Efficacy, U.S. Environmental Protection Agency, Office of Research
and Development, National Homeland Security Research Center, Washington, DC, US EPA
600/R/12/068. 2012.
12	Snedecor, G.W. and Cochran, W.G (1980), Statistical Methods, Seventh Edition, Ames: Iowa State
University Press
13	Tukey, J.W. (1953), "The Problem of Multiple Comparisons," in H.I. Braun, ed., The Collected Works of
John l/l/. Tukey, Volume 8, 1994, New York: Chapman & Hall
14	SAS Institute Inc., 2012, Cary, North Carolina
32

-------
&EPA
United States
Environmental Protection
Agency
PRESORTED STANDARD
POSTAGE & FEES PAID
EPA
PERMIT NO. G-35
Office of Research and Development (8101R)
Washington, DC 20460
Official Business
Penalty for Private Use
$300

-------