EPA/600/R-16/116 I September 2016
www.epa.gov/homeland-security-research
United States
Environmental Protection
Agency
oEPA
Analytical Protocol for VX Using Gas
Chromatography/Mass Spectrometry
(GC/MS)
Office of Research and Development
Homeland Security Research Program

-------
This page left intentionally blank

-------
EPA/600/R-16/116
September 2016
Analytical Protocol for VX Using
Gas Chromatography/Mass
Spectrometry (GC/MS)
United States Environmental Protection Agency
Office of Research and Development
Homeland Security Research Program
Cincinnati, Ohio 45268

-------
Acknowledgments
This method is based on procedures developed by Lawrence Livermore National Laboratory (LLNL)
under Interagency Agreement (IAG) DW89922616-01-0 with the U.S. Environmental Protection Agency
(EPA). EPA's Homeland Security Research Program (HSRP) and Office of Emergency Management
managed and funded multiple-laboratory testing of the procedures for analysis of water, soil, and wipe
samples in a multi-laboratory study. Laboratories participating in the study and providing technical
support include EPA Regions 1, 3, 6, 9 and 10; EPA's Portable High Throughput Integrated Laboratory
Identification System (PHILIS) Unit mobile laboratory in Castle Rock, Colorado; the Virginia Division of
Consolidated Laboratories; the Florida Department of Environmental Protection; and LLNL. Technical
support and data evaluation were provided by CSGov (formerly CSC).
Disclaimer
The U.S. Environmental Protection Agency through its Office of Research and Development funded and
managed the research described herein under EPA Contract No. EP-C-10-060 to CSGov (formerly CSC).
It has been reviewed by the Agency but does not necessarily reflect the Agency's views. No official
endorsement should be inferred. EPA does not endorse the purchase or sale of any commercial products
or services.
Questions concerning this document or its application should be addressed to:
Romy Campisano
U.S. Environmental Protection Agency
Office of Research and Development
National Homeland Security Research Center (NG16)
26 West Martin Luther King Drive
Cincinnati, OH 45268
(513) 569-7016
campisano.romv@,epa.gov

-------
Analytical Protocol for VX Using GC/MS
Table of Contents
Section	Page
I.0	SCOPE AND APPLICATION	1
2.0 SUMMARY OF PROTOCOL	1
3.0 ACRONYMS, ABBREVIATIONS AND DEFINITIONS	1
3.1	Acronyms and Abbreviations	1
3.2	Definitions	2
4.0 INTERFERENCES	5
5.0 SAFETY	5
6.0 EQUIPMENT AND SUPPLIES	6
6.1	General Equipment	6
6.2	Microscale Extraction Apparatus	7
6.3	Gas Chromatograph/Mass Spectrometer (GC/MS) System	8
7.0 REAGENTS AND STANDARDS	10
7.1	Reagents	10
7.2	Standards	10
8.0 SAMPLE PRESERVATION, STORAGE, AND TECHNICAL HOLDING TIMES	12
8.1	Sample Preservation	12
8.2	Sample Storage	12
8.3	Sample Extract Storage	12
8.4	Technical Holding Times	13
9.0 QUALITY CONTROL (QC)	13
9.1	Initial Demonstration of Capability (IDC)	14
9.2	Initial Precision and Recovery (IPR) Determination	14
9.3	Method Blanks	15
9.4	Matrix Spike and Matrix Spike Duplicate (MS/MSD)	17
9.5	Laboratory Control Sample (LCS)	19
9.6	Instrument Detection Limit (IDL) Determination	20
9.7	Method Detection Limit (MDL) Determination	21
9.8	Quantitation Limit (QL) Determination	21
10.0 CALIBRATION AND STANDARDIZATION	22
10.1	Instrument Operating Conditions	22
10.2	GC/MS Mass Calibration (Tuning) and Ion Abundance	23
10.3	Initial Calibration	25
10.4	Continuing Calibration Verification (CCV)	27
10.5	Instrument Blank	29
II.0	ANALYTICAL PROCEDURE	30
11.1	Sample Preparation - General	30
11.2	Preparation of Water Samples Using Microscale Extraction	30
11.3	Preparation of Solid Samples Using Microscale Extraction	31
11.4	Preparation of Wipe Samples by Microscale Extraction	33
11.5	Final Concentration of Extract - Nitrogen Evaporation Technique for solid and wipe
SAMPLES	34
11.6	Extract Analysis by GC/MS	34
/

-------
Analytical Protocol for VX Using GC/MS
12.0 CALCULATIONS AND DATA ANALYSIS	35
12.1	Qualitative Identification of Target Compounds	35
12.2	Data Analysis and Calculations of Target Compounds	36
12.3	Technical Acceptance Criteria for Sample Analysis	40
12.4	Corrective Action for Sample Analysis	41
13.0	ANALYTICAL PROCEDURE PERFORMANCE	42
14.0	POLLUTION PREVENTION	42
15.0	WASTE MANAGEMENT	43
16.0	REFERENCES	43
17.0	TABLES AND FIGURES	46
LIST OF TABLES
Table 1. Instrument Detection Limits (IDL) and Method Detection Limits (MDL) Based on Multi-
Laboratory Evaluation	46
Table 2a. Example Multi-Laboratory Results for Reference Matrix Samples Analyzed Using GC-
Quadrupole MS	47
Table 2b. Example Multi-Laboratory Results for Reference Matrix Samples Analyzed Using GC-TOF
	47
Table 3. Decafluorotriphenylphosphine (DFTPP) Key Ions and Ion Abundance Criteria	48
Table 4. Internal Standard and Surrogates	48
Table 5. Example Retention Times, Relative Retention Times and Quantitation Ions for Target
Compounds, Surrogate Compounds, and Internal Standard	48
Table 6. Example Multi-Laboratory Precision and Bias in Reference Matrices at Mid-Calibration
Levels	49
Table 7. Surrogate Recovery	49
Table 8. Example Calibration Standard Concentrations ((.ig/niL) Used During Multi-Laboratory
Method Validation Study	50
Table 9a. Example Multi-Laboratory Precision and Recovery in Water - Using GC-full scan	51
Table 9b. Example Multi-Laboratory Precision and Recovery in Water - Using GC-TOF	51
Table 10a. Example Multi-Laboratory Precision and Recovery in Soils - Using GC-full scan	52
Table 10b. Example Multi-Laboratory Precision and Recovery in Soils - Using GC-TOF	52
Table 11. Multi-Laboratory Study Water Matrices Characterization Data	53
Table 12. Multi-laboratory Study Soil Matrix Characterization Data	53
LIST OF FIGURES
Figure 1. Example chromatogram for a calibration standard on full-scan quadrupole MS	54
Figure 2. Example chromatogram for a midpoint calibration standard (Cal5) on time-of-flight MS	55
ii

-------
Analytical Protocol for VX Using GC/MS
1.0 SCOPE AND APPLICATION
1.1 The U.S. Environmental Protection Agency (EPA) Homeland Security Research Program
(HSRP) and Office of Emergency Management (OEM), in collaboration with experts
from across EPA and other federal agencies, have identified analytical methods to be
used for the analysis of extractable semi-volatile chemical agents in response to a
homeland security incident. This protocol is to be applied by the national network of
laboratories that has been recruited to the EPA-established Environmental Response
Laboratory Network (ERLN) so that their analytical results are consistent and
comparable. Summaries of these methods are provided in EPA's Selected Analytical
Methods for Environmental Remediation and Recovery (SAM) (Reference 16.1). HSRP
is using the SAM methods (based on the methods listed in Section 1.2) to develop
analytical protocols for laboratory identification and measurement of target agents during
site remediation.
1.2 This document is for the determination of the chemical warfare agent 0-ethyl-S-(2-
diisopropylaminoethyl)methyl-phosphonothiolate (VX). The protocol is based on EPA
Methods 8270D, 3511, 3570, and 1613 (References 16.2, 16.3, 16.4 and 16.5), and a
journal article (Reference 16.6). The procedures have been multi-laboratory tested for
preparation and analysis of VX in soil, wipe, and water samples.
Contaminant

0-ethyl-S-(2-diisopropylaminoethyl)methyl-phosphonothiolate (VX)
50782-69-9
* Chemical Abstracts Service (CAS) Registry Number
1.3 Procedures in this protocol have been tested for the target analyte listed in Section 1.2 in
reference matrices (i.e., reagent water, Ottawa sand, and wipes), drinking water from two
sources, and two dried and homogenized soils. Results of laboratory testing are provided
in Sections 13.0 and 17.0.
2.0 SUMMARY OF PROTOCOL
This analytical protocol involves microscale extraction, followed by gas chromatography/mass
spectrometry (GC/MS) analysis to identify and measure VX. Soil and wipe extracts also might
require a concentration step using nitrogen evaporation to achieve appropriate levels of
quantitation.
3.0 ACRONYMS, ABBREVIATIONS and DEFINITIONS
3.1 Acronyms and Abbreviations
%Recovery Percent recovery
%RSD	Percent relative standard deviation
amu	Atomic mass unit
ASTM	ASTM International (formerly American Society for Testing and
Materials)
CAS RN Chemical Abstracts Service Registry Number
1
September 2016

-------
Analytical Protocol for VX Using GC/MS
ccv
DCM
DF
DFTPP
EI
EICP
EPA
FC-43
GC/MS
HSRP
IDC
IDL
IPR
IS
LCS
MDL
MS/MSD
MSDS
OEM
OSHA
OW
PD
PE
PPE
PTFE
QC
QL
RPD
rpm
RRF
RRT
RSD
RT
SAM
S:N
svoc
TOF
VOA
VX
Continuing calibration verification
Methylene chloride (dichloromethane)
Dilution factor
Decafluorotriphenylphosphine
Electron ionization
Extracted ion current profile
U.S. Environmental Protection Agency
Perfluoro-tri-n-butylamine
Gas chromatograph/mass spectrometer (gas chromatography/mass
spectrometry)
Homeland Security Research Program
Initial demonstration of capability
Instrument detection limit
Initial precision and recovery
Internal standard
Laboratory control sample
Method detection limit
Matrix spike/matrix spike duplicate
Material Safety Data Sheet
Office of Emergency Management
U.S. Occupational Safety and Health Administration
Office of Water
Percent drift
Performance evaluation
Personal protective equipment
Polytetrafluoroethylene (Teflon®)
Quality control
Quantitation limit
Relative percent difference
revolution(s) per minute
Relative response factor
Relative retention time
Relative standard deviation
Retention time
Selected Analytical Methods for Environmental Remediation and
Recovery (SAM), https://www.epa.gov/homeland-securitv-
research/sam (accessed 02/04/2015)
Signal-to-noise ratio
Semivolatile organic compound
Time of flight
Volatile organic analysis
0-ethyl-S-(2-diisopropylaminoethyl)methyl-phosphonothiolate
3.2
Definitions
Aliquot - A measured portion of a field sample, standard, or solution taken for sample
preparation and/or analysis.
Analytical Batch - A set of samples that is analyzed on the same instrument during a 24-
hour period of operation or the analysis of 20 samples (whichever comes first). The
2
September 2016

-------
Analytical Protocol for VX Using GC/MS
analytical batch begins and ends with the analysis of the appropriate Continuing
Calibration Verification (CCV) standards.
Calibration Standard - A solution prepared from the stock standard solution(s) and the
internal standards and surrogate analytes. The calibration standards are used to calibrate
instrument response with respect to analyte concentration.
Continuing Calibration Verification (CCV) - A calibration standard containing the
target analytes, which is analyzed periodically to verify the accuracy of the existing
calibration for those analytes.
Extracted Ion Current Profile (EICP) - A plot of ion abundance versus time (or scan
number) for ion(s) of specified mass(es).
Holding Time - The time elapsed from sample collection until sample extraction or
analysis.
Initial Calibration - Analysis of calibration standards for a series of different specified
concentrations; used to define the quantitative response, linearity, and dynamic range of
the instrument for target analytes.
Initial Demonstration of Capability (IDC) - The IDC is performed prior to use of the
analytical procedures and is used to evaluate the capability of the laboratory in terms of
analytical precision, bias and sensitivity pertaining to the target analytes.
Initial Precision and Recovery (IPR) - A set of four aliquots of a clean reference
matrix (i.e., reagent water, Ottawa sand, clean wipe) to which known quantities of the
target analytes are added. The IPR aliquots are processed and analyzed exactly like a
sample and analyzed prior to the analysis of field samples as part of the IDC. Their
purpose is to determine whether the laboratory is capable of making accurate and precise
measurements.
Instrument Detection Limit (IDL) - The minimum concentration of an analyte that,
when injected into the gas chromatograph/mass spectrometer (GC/MS), produces an
average signal-to-noise ratio (S:N) between 3:1 and 5:1 for at least three replicate
injections.
Instrument Performance Check Solution - A solution of one or more instrument
tuning compounds used to evaluate the performance of the instrument system with
respect to a defined set of method criteria.
Internal Standard (IS) - Analyte added to an extract or standard solution in a known
amount and used to measure the relative responses of target analytes and surrogates. The
internal standard must be an analyte that is not a sample component.
Laboratory Control Sample (LCS) - An aliquot of a clean reference matrix (i.e.,
reagent water, Ottawa sand, clean wipe) to which known quantities of the target analytes
are added. The LCS is processed and analyzed exactly like a sample. Its purpose is to
determine whether the analytical process is in control.
3
September 2016

-------
Analytical Protocol for VX Using GC/MS
Safety Data Sheet (SDS) - Written information provided by vendors concerning a
chemical's toxicity, health hazards, physical properties, flammability, and reactivity data
including storage, spill, and handling precautions.
Matrix - The predominant material of which the sample to be analyzed is composed.
For the purpose of this protocol, a sample matrix is either aqueous/water,
soil/sediment/sand, or wipe. Matrix is not synonymous with phase (e.g., liquid or solid).
Matrix Spike (MS) - An aliquot of a field sample to which known quantities of target
analytes are added. The MS is processed and analyzed exactly like the corresponding
sample. Its purpose is to determine whether the sample matrix contributes bias to the
analytical results. Background concentrations of the analytes must be determined in a
separate aliquot.
Matrix Spike Duplicate (MSD) - A second aliquot of the field sample used to prepare
the MS, which is fortified, extracted and analyzed exactly like the MS. The MSD is used
to assess matrix effects on analytical precision and bias.
Method Blank - An aliquot of a clean reference matrix (i.e., reagent water, Ottawa sand,
clean wipe) that is treated exactly as a sample, including exposure to all glassware,
equipment, solvents, reagents, sample preservatives, internal standards, and surrogates
that are used in the extraction batch. The method blank is used to determine whether
target analytes or interferences are present in the laboratory environment, reagents or
equipment.
Method Detection Limit (MDL) - The minimum concentration of an analyte that can be
identified, measured and reported with 99 percent confidence that the analyte
concentration is greater than zero. The MDL is a statistical determination (Section 9.7),
and accurate quantitation at this level is not expected.
Percent Difference - The difference between two values divided by one of the values.
Used in this protocol to compare two relative response factor (RRF) values from
calibration.
Percent Drift (PD) - The difference between the calculated and theoretical value divided
by the theoretical value. Used in this protocol to compare calculated and theoretical
values for calibration by regression techniques.
Quantitation Limit (QL) - The minimum level of quantitation. This concentration must
meet the criteria defined in Section 9.8.
Reagent Water - Water in which an interferent is not observed at or above the low-level
calibration standard for each analyte of interest. The purity of this water must be
equivalent to ASTM International (ASTM) Type II reagent water of Specification
D1193-06, "Standard Specification for Reagent Water."
Relative Percent Difference (RPD) - The difference between two values divided by the
mean of the values. RPD is reported as an absolute value (i.e., always expressed as a
positive number or zero).
4
September 2016

-------
Analytical Protocol for VX Using GC/MS
Relative Response Factor (RRF) - A measure of the relative mass spectral response of
an analyte compared to its internal standard. RRFs are determined by analysis of
standards and are used in calculating the concentrations of analytes in samples.
Retention Time (RT) - The time an analyte is retained on a GC column before elution.
The RT is dependent on the analyte, nature of the column's stationary phase, the
column's diameter, temperature, carrier gas flow rate, and other column parameters.
Relative Retention Time (RRT) - The ratio of the RT of a compound to the RT of a
corresponding internal standard.
Stock Standard Solution - A concentrated solution containing one or more target
analytes prepared in the laboratory using assayed reference materials or materials
purchased from a reputable commercial source.
Surrogate - Analyte that is unlikely to be found in any sample to be analyzed.
Surrogates are added to a sample aliquot in a known amount before extraction or other
processing. Surrogates are measured with the same procedures used to measure other
sample components. The purpose of the surrogate is to monitor method performance
with each sample.
Working Standard Solution - A solution containing target analytes prepared from stock
standard solutions. Working standard solutions are diluted as needed to prepare
calibration and spiking solutions.
4.0 INTERFERENCES
4.1	Contaminants in solvents, reagents, glassware, and other sample processing hardware can
cause method interferences such as discrete artifacts and/or elevated baselines in the
extracted ion current profiles (EICPs). All of these materials must be routinely
demonstrated to be free from interferences under the conditions of the analysis by
running method blanks. Matrix interferences can be caused by contaminants that are co-
extracted from the sample. The extent of matrix interferences will vary considerably
from source to source and is evaluated using the results from the analysis of matrix spike
and matrix spike duplicate samples. In addition to the interferences mentioned above,
VX is subject to matrix enhancement effects reported frequently for gas chromatographic
analysis of various compounds, potentially resulting in high unexpected recoveries in
some samples.
4.2	This protocol includes conditions for collecting mass spectral data using both quadrupole
mass spectrometers in full-scan mode and time-of-flight (TOF) mass spectrometers.
5.0 SAFETY
WARNING: The toxicity of VX presents hazards unfamiliar to most experienced laboratory
personnel. Special techniques and precautions must be used even for the simplest procedures
involving these agents. Because VX is the target analyte for this protocol, laboratory personnel
must be thoroughly trained in appropriate safety procedures prior to using this method.
5
September 2016

-------
Analytical Protocol for VX Using GC/MS
5.1	Operations with VX have specific safety requirements. The laboratory must have these
requirements included in a Chemical Hygiene Plan prior to conducting the analytical
procedures described in this protocol.
5.2	At a minimum, personal protective equipment (PPE) requirements include safety glasses,
lab coats, and protective gloves. The availability of emergency response equipment and
support personnel should be as indicated in a laboratory Chemical Hygiene Plan.
5.3	Exposure to chemical agent material is possible from contact, and risk is primarily
associated with compromise of PPE. Respiratory exposure can result from spills or
improper use of ventilation controls and PPE.
5.4	At concentrations of VX that are within the calibration range of this method, the likelihood
of an exposure causing adverse health effects is extremely low (Reference 16.7).
Note: This method does not address all safety issues associated with its use. The
laboratory is responsible for maintaining a safe work environment and a current file of
Occupational Safety and Health Administration (OSHA) regulations regarding the safe
handling of the chemicals (including all solvents, reagents, and standards). A reference
file of safety data sheets (SDSs) must be available to all personnel involved in these
analyses, chemical handling, and contaminated area cleaning, or who might potentially
come in contact with the materials in their work place.
6.0 EQUIPMENT AND SUPPLIES
Brand names, suppliers, catalog, and part numbers are for illustrative purposes only. No
endorsement is implied. Equivalent performance can be achieved using equipment and supplies
other than those specified; however, laboratories must document use of the alternatives and
provide a demonstration of equivalent performance meeting the requirements of this protocol.
6.1 General Equipment
6.1.1	Vials - Clear or amber glass, with polytetrafluoroethylene (PTFE)-lined screw
or crimp top (2.0-mL capacity for GC auto sampler) (Sigma Aldrich Catalog
No.SU860033, Sigma-Aldrich. St. Louis, MO) or equivalent. Glass inserts can
be used to minimize sample volumes.
6.1.2	Syringes - Contaminant-free, 1.0 mL, 2.0 mL, 10 |iL. 25 (jL, 500 jjL
6.1.3	Pasteur glass pipettes - 1.0 mL, disposable (Fisher Scientific Catalog No.
NC0541803, Thermo Fisher Scientific, Westminster, MD) or equivalent
6.1.4	Balances - Analytical, capable of accurately weighing ±0.0001 gram, and one
capable of weighing 100 grams (±0.01 gram).
6.1.5	Spatula - Stainless steel or PTFE
6.1.6	Nitrogen evaporation device - Equipped with temperature control that can be
maintained at 35 - 40°C, a RapidVap® evaporator (Labconco Corporation,
Kansas City, MO) or equivalent. To prevent the release of solvent fumes into
6
September 2016

-------
Analytical Protocol for VX Using GC/MS
the laboratory, this device must be used with suitable engineering controls.
Water bath (for nitrogen evaporation devices that do not have a heating
apparatus) - Heated, with concentric ring cover capable of heating to 80 °C and
maintaining temperature control (±5 °C). The bath should be used with suitable
engineering controls.
Glass funnel (Fisher Scientific Catalog No. CG172305, Thermo Fisher
Scientific, Westminster, MD) or equivalent - Used in filtering soil samples that
fail to settle out with centrifugation.
Borosilicate glass wool - Oven-cleaned (muffled) or solvent rinsed (using
extraction solvent); used in filtering soil samples that fail to settle out with
centrifugation.
pH paper - Including narrow range capable of measuring a pH of 2.0.
pH meter - With a combination glass electrode. Calibrate prior to each use
according to manufacturer's instructions.
Ottawa sand - Held at 450 °C for four hours in a 500-mL wide-mouthed amber
bottle.
Vortexer - VWR™ vortexer (VWR Corporation, Radnor, PA) or equivalent,
capable of accommodating 40 - 60-mL vials and 50-mL centrifuge tubes.
Shaker table
6.1.14.1	Glas-Col Large Capacity Mixer (Part # 099A LC1012, Glas-Col Inc.,
Terre Haute, IN); or Glas-Col Digital Pulse Mixer (Part # 099A
DPMI2) or equivalent
6.1.14.2	Foam pad for 40-mL volatile organic analysis (VOA) vials (Part
#099A VC4014, Glas-Col Inc., Terre Haute, IN or equivalent)
6.1.14.3	Foam pad for 60-mL VOA vials (Part #099A VC6014, Glas-Col
Inc., Terre Haute, IN) or equivalent
6.2 Microscale Extraction Apparatus
6.2.1 Solid samples
6.2.1.1	VOA vials - 40 mL capacity, disposable, pre-cleaned with PTFE-
lined caps (Fisher Scientific Catalog No. 05-719-400,Thermo Fisher
Scientific, Westminster, MD or equivalent). If pre-cleaned vials are
not available, appropriate cleaning procedures are provided in EPA
Methods 525.3 (Reference 16.8) and 1668C (Reference 16.9), and in
SW-846 Chapter 4 (Reference 16.10).
6.2.1.2	Sonicator - Branson 3510 sonicator (Branson Ultrasonics Corp.,
Danbury, CT) or equivalent
6.1.7
6.1.8
6.1.9
6.1.10
6.1.11
6.1.12
6.1.13
6.1.14
7
September 2016

-------
Analytical Protocol for VX Using GC/MS
6.2.1.3	Apparatus for determining percent dry weight
6.2.1.3.1	Drying oven - Capable of maintaining 103 - 105 °C
6.2.1.3.2	Desiccator
6.2.1.3.3	Crucibles - Disposable aluminum
6.2.1.4	Glass beads - Solvent-rinsed, baked in 400 °C oven for
approximately 1 hour (Fisher Scientific Catalog No. S80024 or
equivalent)
6.2.1.5	Centrifuge - Capable of at least 500 G-force units and
accommodating 40- or 50-mL vials, Accuspin™ Model 400
centrifuge (Thermo Fisher Scientific, Westminster, MD) or
equivalent. CAUTION: Different centrifuge makes and models have
different maximum centrifuge speeds that are recommended for safe
operation. The maximum safe handling speed of each centrifuge will
depend, in part, on the vials used and should be determined prior to
centrifuging samples.
6.2.1.6	Pasteur pipettes - 1.0-mL glass, disposable orre-
pipettes/autopipettes with disposal tips
6.2.2	Water Samples
6.2.2.1	Conical bottom glass screw-top tube, 50-mL or 60-mL vials, pre-
cleaned with PTFE-lined caps. If pre-cleaned vials are not available,
appropriate cleaning procedures are provided in EPA Methods 525.3
(Reference 16.8) and 1668C (Reference 16.9), and in SW-846
Chapter 4 (Reference 16.10).
6.2.2.2	Beakers - 400 mL
6.2.2.3	Class A graduated cylinder - 100 mL
6.2.2.4	Class A volumetric flasks - 10 mL
6.2.3	Wipe samples - Kendall Curity® gauze sponges (USP Type VII Gauze, cotton,
12 ply 3 in x 3 in, Tyco Healthcare Group, Covidien, Mansfield, MA) or
equivalent; pre-cleaned by extracting with methylene chloride (DCM) using an
appropriate extraction method (e.g., pressurized fluid extraction, Soxhlet
extraction, soaking in DCM).
Note: Wipes used for field or laboratory quality control (QC) samples should be
pre-wetted with DCM prior to use.
6.3 Gas Chromatograph/Mass Spectrometer (GC/MS) System
6.3.1 Gas chromatograph - The GC system must be capable of temperature
programming and have a flow controller that maintains a constant column carrier
8
September 2016

-------
Analytical Protocol for VX Using GC/MS
gas flow rate throughout the temperature program operations. The system must
be suitable for splitless injection and have all required accessories including
syringes, analytical columns, and gases. All GC carrier gas lines must be
constructed from stainless steel or copper tubing. Non-PTFE thread sealants or
flow controllers with rubber components are not to be used.
Note: Due to the potential hazards associated with the analysis of VX, it is
recommended that a split vent trap be used for every GC system (Agilent® RD-
1020 Universal/External Split Vent Trap (Agilent Technologies, Inc., Santa
Clara, CA or equivalent). For additional safety measures, the split and purge
vent lines can be vented to the hood ventilation system.
6.3.2	GC column - Recommended length 30 m x 0.25 mm internal diameter (ID) (or
0.32 mm) bonded phase silicon coated fused silica capillary column DB-5 (J&W
Scientific, Agilent Technologies, Santa Clara, CA); DB-5MS; RTX®-5 (Restek
Corp., Bellefonte, PA), RTX-5MS (Restek Corp., Bellefonte, PA), RTX-5Sil MS
(Restek Corp., Bellefonte, PA); Zebron® ZB-5 (Phenomenex®, Phenomenex Inc.,
Torrance, CA); SPB-5 (Supelco®, Sigma-Aldrich. St. Louis, MO); AT™-5
(Alltech®, Grace, Columbia, MD); HP-5 (Agilent Technologies, Santa Clara,
CA); HP-5MS or HP-5MS UI (Agilent Technologies, Santa Clara, CA), CP-Sil 8
CB (Chrompack, Raritan, NJ); 007-2 (Quadrex®, Quadrex Corp., Bethany, CT);
BP-5 (SGE, Trajan Scientific Americas, Inc., Austin, TX); Zebron ZB-5MS
(Phenomenex®, Phenomenex Inc., Torrance, CA) or equivalent. Columns used to
generate the example data provided in Section 17 include: Agilent HP-5 MS,
RTX - 5sil, RTX-5MS, RTX-5Sil MS, Zebron ZB-5 MS, DB-5MS.
Although a film thickness of 1.0 micron might be desirable because of its larger
capacity, film thicknesses of 0.25 micron and 0.05 micron were used by
laboratories generating the example data presented in Section 17. A description
of the GC column used for analysis shall be provided in the data narrative. A
capillary column is considered equivalent if:
•	The column does not introduce contaminants that interfere with the
identification and quantification of the compounds listed in Table 1, Section
17.
•	The analytical results generated using the column meet the initial CCV
technical acceptance criteria listed in the protocol and the quantitation levels
determined as described in Section 9.8.
•	The column provides equal or better resolution of the compounds listed in
Table 1, Section 17 when compared to columns listed above.
6.3.3	MS - Must be capable of recording a spectrum from 35 - 500 atomic mass units
(amu) every second or less, using 70 electron volts (nominal) in the electron
ionization (EI) mode, and producing a mass spectrum that meets the tuning
acceptance criteria when 50 ng or less of decafluorotriphenylphosphine
(DFTPP) is injected through the GC inlet. The instrument must be vented to
prevent the release of contaminants into the instrument room.
9
September 2016

-------
Analytical Protocol for VX Using GC/MS
7.0 REAGENTS AND STANDARDS
7.1	Reagents
7.1.1	Organic-free reagent water - Water in which an interferent is not observed at or
above the quantitation limit (QL) for each analyte of interest. ASTM Type II
reagent water of Specification D1193-06, "Standard Specification for Reagent
Water" (Reference 16.11) or equivalent.
7.1.2	Acetone, DCM, and toluene - pesticide residue analysis grade or equivalent.
DCM must contain a non-methanol olefinic stabilizer (i.e., amylene or 2-
pentene). Solvents must be anhydrous or pre-dried by adding sodium sulfate
until no clumping occurs and the solution becomes clear (not cloudy).
7.1.3	Sodium sulfate - Powdered or granular anhydrous reagent grade, heat at 400°C
for four hours in a shallow tray, cool in a desiccator, and store in a glass bottle.
7.1.4	Tris-buffer (for soil extraction of VX) - 2 L of Tris-buffer is prepared as a
solution containing 0.2 M tris(hydroxymethyl)aminomethane (Trizma® base,
Product # T1503, Sigma Aldrich or equivalent, Sigma-Aldrich, St. Louis,
MO) and 32 mL of 1 M hydrochloric acid (HC1). The pH of the solution should
be in the range of 8.5 - 9.0. If necessary, adjust accordingly with additional
tris(hydroxymethyl)aminomethane to raise the pH, or HC1 to lower it.
7.1.5	Sodium chloride - Anhydrous reagent grade, >99 %.
7.1.6	Sodium thiosulfate - Reagent grade, 10 % sodium thiosulfate solution in reagent
water. Prepared fresh with each use.
7.2	Standards
The laboratory must be able to verify that stock standard solutions are certified.
Manufacturers' certificates of analysis must be retained by the laboratory and presented
upon request. Stock standard solutions provided in sealed glass ampules may be retained
and used within six months of the preparation date. Solutions used for calibration
verification ideally are prepared from a separate source other than the source used to
prepare calibration standards. Due to the nature of the analyte addressed in this protocol,
identification of a secondary source might be difficult.
7.2.1	Stock standard solutions
Stock standard solutions, used to produce working standards, may contain
individual target compounds or mixtures of target compounds.
7.2.2	Working standards
7.2.2.1 Surrogate standard spiking solution - Prepare a surrogate standard
spiking solution in DCM or other suitable solvent that contains
appropriate surrogates for the target compound. A concentration of
25 (ig/mL is recommended for each surrogate. Surrogate standards
are added to all samples and calibration solutions.
10
September 2016

-------
Analytical Protocol for VX Using GC/MS
Note: Table 4 provides a list of surrogates used during laboratory
studies testing this protocol, based on surrogates typically used for
semivolatile organic compounds (SVOC) analyses. Alternative
surrogates might be more representative of the analyte targeted in
this method, and may be used instead of or in addition to those listed
in Table 4, Section 17, provided the surrogates meet the criteria in
Table 7, Section 17.
7.2.2.2	Matrix spiking solution - This solution is prepared in DCM and
should contain the target analyte.
7.2.2.3	Instrument performance check solution - Prepare a solution of
DFTPP in DCM such that a 1 -|_iL injection will contain 50 ng or less
of DFTPP. The DFTPP may also be included in the calibration
standards at this level.
7.2.2.4	Initial and continuing calibration solutions
7.2.2.4.1 Prepare calibration standards in DCM at a minimum of
five concentration levels. Each calibration standard
should contain VX, associated surrogates, and internal
standard.
Note 1: All samples analyzed must be injected at the
same volume (e.g., 1.0 or 2.0 (.iL) as the calibration
standards.
Note 2: The concentrations listed in Table 8, Section 17,
provide an example calibration range used during
laboratory evaluation of this protocol. The low
calibration standard is set at the expected QL
(determined in Section 9.8). The remaining calibration
standards should be prepared at concentrations that meet
the specifications in Section 10.3.4.
7.2.2.4.2 The CCV standard is prepared in DCM at or near the
midpoint of the calibration curve.
7.2.2.5 Internal standard solution - An internal standard solution can be
prepared by dissolving 100 mg of phenanthrene-dio in 100 mL of
DCM, resulting in a concentration of 1.0 mg/mL. A sufficient portion
of this solution will be added to each sample extract just prior to
analysis by GC/MS to result in a concentration of 0.5 ng/(.iL (for both
full-scan quadrupole and TOF mass spectrometers). Alternatively,
internal standard solutions can be purchased from commercial sources
(e.g., Supelco part numbers 861238 or 48710-U, or equivalent,
[Supelco®, Sigma-Aldrich. St. Louis, MO]).
11
September 2016

-------
Analytical Protocol for VX Using GC/MS
7.2.3 Storage of standard solutions
7.2.3.1	Store unopened ampules of stock standard solutions at <6 °C. If no
manufacturer's expiration date is provided, unopened ampules of
standard solutions may be retained and used for up to twelve months
after the preparation date (Reference 16.12). Store opened stock
standard solutions at <6 °C in PTFE-lined screw-cap amber bottles.
Opened stock standards containing only VX are stable for up to 6
months; opened stock standards containing chemical warfare agents
in addition to VX (i.e., sarin, soman, cyclohexyl sarin and/or sulfur
mustard) are stable for up to six days.
7.2.3.2	Store the working standards at < 6 °C in containers with PTFE-lined
caps. Working standard solutions should be checked against CCV
standards at least weekly for stability, and must be replaced after six
months, or sooner, if the stock standard solutions have expired or if
comparison with CCV samples indicates a problem.
7.2.3.3	Protect all standards from light. Samples, sample extracts, and
standards must be stored separately.
7.2.3.4	The laboratory is responsible for maintaining the integrity of standard
solutions and verifying the solutions prior to use. The standards must
be brought to room temperature prior to use, checked for losses, and
checked to ensure that all components have remained in solution.
Guidance on standard verification procedures can be found in EPA's
Superfund Analytical Services / Contract Laboratory Program, Multi-
Media, Multi-Concentration Organics Analysis, SOM02.3, Exhibit E,
Section 4 (Reference 16.13).
8.0 SAMPLE PRESERVATION, STORAGE, AND TECHNICAL HOLDING TIMES
8.1	Sample Preservation
Samples must be stored on ice or refrigerated at 4 °C (± 2 °C) immediately after
collection until receipt in the laboratory. If chlorine is suspected to be present in water
samples (e.g., treated drinking water or wastewater) that are to be measured for VX, add
approximately four drops (-0.2 mL) of a 10 % solution of sodium thiosulfate per 35-mL
sample. If clouding results, add less sodium thiosulfate to a fresh sample aliquot. If
sodium thiosulfate is not added during sample collection, it should be added immediately
upon sample receipt in the laboratory, prior to sample analysis or extraction.
8.2	Sample Storage
Samples must be protected from light and refrigerated at 4 °C (± 2 °C) from the time of
receipt until extraction.
8.3	Sample Extract Storage
8.3.1 Sample extracts must be protected from light and stored at <6 °C.
12
September 2016

-------
Analytical Protocol for VX Using GC/MS
8.3.2 Samples, sample extracts, and standards must be stored separately.
8.4 Technical Holding Times
8.4.1 Soil and wipe samples must be extracted within seven days from receipt at the
laboratory. Aqueous samples must be extracted within 14 days from receipt at
the laboratory.
Note: Holding times for soils and wipes are based on holding times listed in
guidance documents and similar analytical methods for SVOCs (e.g., EPA
Methods 525.2 [Reference 16.14] and 525.3 [Reference 16.8], CLP Method
SOM 2.3 [Reference 16.15], SW-846 Chapter 4 [Reference 16.10]). Holding
times for water samples were evaluated in a single laboratory using this protocol.
8.4.2 Extracts must be analyzed within 14 days following extraction.
9.0 QUALITY CONTROL (QC)
QC requirements for this protocol include the following:
Quality Control (QC) Analyses
Requirement
Section
Frequency
Instrument Detection Limit (IDL)
Determination
Section 9.6
Optional. Performed prior to Method
Detection Limit (MDL) study
Method Detection Limit (MDL)
Determination
Section 9.7
Performed once, prior to first performing
the protocol procedures and with each
significant change as part of the Initial
Demonstration of Capability (IDC)
Initial Precision and Recovery (IPR)
Determination
Section 9.2
Quantitation Limit (QL) Determination
Section 9.8
Method Blanks
Section 9.3
At least one per batch of <20 samples of
the same matrix
Instrument Blank
Section 10.5
Following an analysis with suspected
carry-over or following analysis of
samples containing high concentrations
Matrix Spike / Matrix Spike Duplicate
(MS/MSD)
Section 9.4
One per each batch of <20 samples of the
same matrix
Laboratory Control Sample (LCS)
Section 9.5
At least one per batch of <20 samples of
the same matrix
Continuing Calibration Verification (CCV)
Section 10.4
Prior to the analysis of samples, and after
instrument performance check. Analyzed
at the beginning and at the end of each
analytical batch of <20 injections
Precision and bias criteria for data generated using this method are currently set at 50 - 150 %
recovery and < 30 % precision (as relative standard deviation [RSD] or relative percent difference
[RPD]). These criteria may change as more laboratory data become available. In cases where
analyses of difficult sample matrices generate results outside these criteria, data should be
flagged, and laboratories should collect additional data to support development of laboratory- and
matrix-specific criteria. Example precision and bias results obtained from multiple laboratories
analyzing spiked reference matrix samples (reagent water, Ottawa sand, and wipes) and field
13
September 2016

-------
Analytical Protocol for VX Using GC/MS
samples (water and soil) are provided in Section 17.
9.1	Initial Demonstration of Capability (IDC)
An IDC shall be performed prior to the analysis of any samples and with each significant
change in instrument type, detection technique, personnel or method. An IDC consists of
the following:
•	An IPR determination (Section 9.2)
•	An MDL determination (Section 9.7)
•	A QL determination (Section 9.8) on a clean matrix (reagent water, Ottawa sand, pre-
cleaned wipe)
The IPR consists of four replicate samples of a clean matrix spiked with VX around the
midpoint of the calibration curve and carried through the entire analytical process. Prior
to performing the IDC, a valid initial calibration (Section 10.3) must be established.
9.2	Initial Precision and Recovery (IPR) Determination
9.2.1 Preparation and analysis of IPR samples
9.2.1.1	Water samples
Prepare four replicate samples from about 35 mL of reagent water.
Add a sufficient amount of surrogate standard spiking solution and
matrix spiking solution to result in the addition of 1.0 (.ig of each
surrogate (add 40 |_iL if prepared as in Section 7.2.2.1) and a
concentration at the mid-point of the calibration range of VX.
Extract and analyze according to the procedures for water samples
(Sections 11.2 and 11.6). The total volume of DCM added will be
slightly greater than the 2 mL needed for extraction, and includes the
volumes added for spiking VX and the surrogate.
9.2.1.2	Ottawa sand
Prepare four replicate samples consisting of 10 grams of Ottawa
sand. Add a sufficient amount of the surrogate standard spiking
solution and the matrix spiking solution to result in the addition of
0.5 |ag of each surrogate (add 20 (.iL if prepared as in Section 7.2.2.1)
and a concentration at the mid-point calibration range of VX, and
follow the appropriate extraction procedure in Section 11.3. Extract,
concentrate and analyze according to procedures for solid samples.
9.2.1.3	Pre-cleaned wipes
Prepare four replicate samples consisting of pre-cleaned wipes. Pre-
wet the wipes with DCM prior to use. Add a sufficient amount of
the surrogate standard spiking solution and the matrix spiking
solution to result in the addition of 0.5 |_ig of each surrogate (add 20
(.iL if prepared as in Section 7.2.2.1) and a concentration at the mid-
point calibration range of VX, and follow the appropriate extraction
procedure in Section 11.4. Extract, concentrate and analyze
14
September 2016

-------
Analytical Protocol for VX Using GC/MS
according to procedures for wipe samples.
9.2.2	Calculations for IPR
9.2.2.1	Calculate the percent recovery (%Recovery) of each compound in the
IPR sample using Eq. 11 (Section 12.2.9.2). Calculate an average
%Recovery for each compound.
9.2.2.2	Calculate a percent relative standard deviation (%RSD) for each
compound in the IPR samples.
9.2.3	Technical acceptance criteria for IPR
9.2.3.1	The average percent recovery of each compound in the IPR should be
within the 50 - 150 %.
9.2.3.2	The %RSD of each compound in the IPR should be less than or equal
to 30.
9.2.4	Corrective action for IPR
If the technical acceptance criteria in Section 9.2.3 are not met, inspect the
system for problems and take corrective actions to achieve the acceptance
criteria.
9.3 Method Blanks
A method blank is a volume of a clean reference matrix (e.g., reagent water for water
samples, clean inert sand for solid samples, or clean pre-wetted wipes for wipe samples)
spiked with a sufficient amount of surrogate standard spiking solution (Section 7.2.2.1)
such that the same amount of surrogate is added as for the associated samples and carried
through the entire analytical procedure. Internal standard solution is added just prior to
analysis by GC/MS to give an internal standard concentration of 0.5 ng/|_iL for both full-
scan and TOF modes. The volume or weight of the method blank must be approximately
equal to the volume or weight of the samples associated with the blank.
9.3.1	Frequency of method blanks
A method blank must be extracted each time samples are extracted. The number
of samples extracted with each method blank should not exceed 20 field samples
(excluding MS/MSDs and Performance Evaluation [PE] samples). In addition, a
method blank is:
•	Extracted by the same procedure used to extract samples
•	Analyzed on each GC/MS system used to analyze associated samples and
conditions (i.e., GC/MS settings)
9.3.2	Method blank preparation
9.3.2.1 A method blank for water samples consists of a 35-mL volume of
reagent water spiked with a sufficient amount of the surrogate
15
September 2016

-------
Analytical Protocol for VX Using GC/MS
standard spiking solution to result in the addition of 1.0 (ig of
surrogate (add 40 |_iL if prepared as in Section 7.2.2.1). The final
volume of the extracts will be approximately 2.0 mL; therefore, the
concentration of the surrogate in the extract is expected to be
approximately 0.50 |_ig/mL. For solid samples, a method blank
consists of 10 grams of clean inert sand spiked with a sufficient
amount of the surrogate spiking solution to result in the addition of
0.5 (.ig of surrogate (add 20 |_iL if prepared as in Section 7.2.2.1). The
final volume of the extracts will be 1.0 mL; therefore, the
concentration of the surrogate in the extract is expected to be 0.50
(ig/mL. A method blank for wipe samples consists of a clean unused
wipe spiked with a sufficient amount of the surrogate standard
spiking solution to result in the addition of 0.5 (ig of surrogate (add
20 |_iL if prepared as in Section 7.2.2.1). The final volume of the
extracts will be 1.0 mL; therefore, the concentration of the surrogate
in the extract is expected to be 0.50 |_ig/mL. Extract, concentrate, and
analyze the blank according to procedures.
9.3.2.2 Under no circumstances should method blanks be analyzed at a
dilution.
9.3.3 Technical acceptance criteria for method blank analysis
9.3.3.1 All blanks should be extracted and analyzed at the frequency
described in Section 9.3.1 on a GC/MS system meeting the DFTPP
tuning criteria in Section 10.2.4 and Table 3, Section 17, initial
calibration in Section 10.3, and CCV technical acceptance criteria in
Section 10.4.5.
9.3.3.2	The %Recovery of each of the surrogates in the blank must be within
the acceptance limits listed in Table 7, Section 17.
9.3.3.3	The blank must meet the sample analysis acceptance criteria listed in
Section 12.3.
9.3.3.4 A method blank for solid, water, and wipe samples must contain a
concentration less than the MDL of VX. In cases where a blank has
detects above the QL, but associated samples have detects greater
than 10 times the blank, consult the agency to determine if re-
extraction is required. In cases where a method blank fails to meet
technical acceptance criteria, but all samples had non-detects for the
target analyte, then no re-extraction or qualification of data is
necessary.
9.3.4 Corrective action for method blanks
9.3.4.1	If a method blank does not meet the technical acceptance criteria for
method blank analysis, the analytical system is considered to be out
of control.
9.3.4.2	If contamination is the problem, then the source of the contamination
16
September 2016

-------
Analytical Protocol for VX Using GC/MS
should be investigated and appropriate corrective measures taken
before further sample analysis proceeds. It is the laboratory's
responsibility to ensure that interferences caused by contaminants in
solvents, reagents, glassware, and sample storage and processing
hardware that lead to discrete artifacts and/or elevated baselines in the
GC/MS have been eliminated. If possible, an aliquot of any sample
associated with the contaminated blank should be re-extracted and re-
analyzed.
9.3.4.3	If surrogate recovery in the method blank does not meet the
acceptance criteria listed in Table 7, first reanalyze the method blank.
If the surrogate recovery does not meet the acceptance criteria after
reanalysis, the method blank and an aliquot of any sample associated
with that method blank should be re-extracted, if possible, and re-
analyzed. If the surrogate recovery is high and all corresponding
samples had non-detects for VX, sample re-extraction and re-analysis
are not required.
9.3.4.4	If the method blank does not meet the internal standard response
requirements in Section 12.3.5, follow the corrective action procedure
in Section 12.4.4.1. Resolve and document problem resolution before
proceeding with sample analysis.
9.3.4.5	If the method blank does not meet the retention time (RT)
requirements for the internal standard (Section 12.3.6), check the
instrument for malfunction and recalibrate. Reanalyze the method
blank.
9.3.4.6	Samples that are analyzed with corresponding method blanks that do
not meet any of the criteria listed in Sections 9.3.4.2 - 9.3.4.5 should
be reanalyzed. If the method does not meet the criteria, then all
corresponding sample data should be flagged.
9.4 Matrix Spike and Matrix Spike Duplicate (MS/MSD)
To evaluate the potential effects of the sample matrix on analyses, VX must be spiked
into two additional aliquots of a water or solid sample and analyzed in accordance with
the appropriate method. VX should be spiked at a concentration near the midpoint of the
calibration range.
9.4.1 Frequency of MS/MSD analyses
9.4.1.1	An MS/MSD pair is analyzed with each batch of <20 samples of each
water or solid matrix type. MS/MSDs are not performed on wipe
samples.
9.4.1.2	For quality assurance purposes, water rinsate samples and/or field
blanks (field QC) or PE samples may accompany solid, water, and/or
wipe samples that are delivered to the laboratory for analysis. These
field QC or PE samples are not used for MS/MSD analyses.
17
September 2016

-------
Analytical Protocol for VX Using GC/MS
9.4.1.3	If the agency requesting the analyses designates a sample to be used
as an MS/MSD, then that sample must be used. If there is insufficient
sample remaining to perform an MS/MSD, then the laboratory shall
choose another sample on which to perform an MS/MSD analysis. At
the time the selection is made, the laboratory should notify the agency
that insufficient sample was received and identify the sample selected
for the MS/MSD analysis.
9.4.1.4	If there is insufficient sample remaining in any of the samples in a
batch to perform the required MS/MSD, the laboratory will report this
in the data narrative.
9.4.2 Procedure for Preparing MS/MSD
9.4.2.1 Water samples
Prepare two additional aliquots of the sample chosen for spiking.
The volume should be equal to that of the associated samples. Add a
sufficient amount of surrogate standard spiking solution and matrix
spiking solution to each aliquot to result in the addition of 1.0 (ig of
each surrogate (add 40 |_iL if prepared as in Section 7.2.2.1) and a
concentration at the mid-point of the calibration range of VX.
Extract, clean up, and analyze the MS/MSD according to the
procedures for water samples (Section 11.2). The total volume of
DCM added will be slightly greater than the 2 mL needed for
extraction, and includes volumes added for spiking VX and the
surrogate.
9.4.2.2 Solid samples
Prepare two additional aliquots of the sample chosen for spiking in
two 40-mL VOA vials with PTFE-lined caps. The amount chosen
should be equal to that of the associated sample.
Mix well. Add a sufficient amount of the surrogate standard spiking
solution and the matrix spiking solution to each aliquot to result in
the addition of 0.5 |_ig of each surrogate (add 20 |_iL if prepared as in
Section 7.2.2.1) and a concentration at the mid-point of VX, and
follow the appropriate extraction procedure in Section 11.3. Extract,
concentrate, clean up, and analyze the MS/MSD according to
procedures for solid samples.
9.4.3 Dilution of MS/MSD
Before any MS/MSD analysis, analyze the original sample, then analyze the
MS/MSD at the same concentration as the most concentrated extract for which
the original sample results will be reported.
9.4.4 Calculations for MS/MSD
9.4.4.1 Calculate the %Recovery of each matrix spike compound in the
18
September 2016

-------
Analytical Protocol for VX Using GC/MS
MS/MSD sample (see Eq. 11 in Section 12.2.9).
9.4.4.2 Calculate the Relative Percent Difference (RPD) of the concentrations
of each compound in the MS/MSD using Eq. 1. Concentrations of
the matrix spike compounds are calculated using the sample equations
used for target compounds (Eq. 6 for water samples and Eq. 7 for
solid samples in Section 12.2.6).
Eq. 1 Relative Percent Difference Calculation
RPD = }C' ~Cl * x 100
ct+c2j
where:
Ci = Measured concentration of the first sample aliquot
C2 = Measured concentration of the second sample aliquot
9.4.5	Technical acceptance criteria for MS/MSD
9.4.5.1	All MS/MSDs must be analyzed on a GC/MS system meeting DFTPP
and initial CCV technical acceptance criteria, as well as the method
blank technical acceptance criteria.
9.4.5.2	The MS/MSD must be extracted and analyzed within the technical
holding time (Section 8.4).
9.4.5.3	The RT shift for the internal standard must be within ± 0.50 minutes
(30 seconds) between the MS/MSD sample and the most recent CCV
standard analysis.
9.4.5.4	The limits for matrix spike compound recovery and RPD are given 50
- 150 % and < 30 % respectively. For difficult matrices, laboratories
are encouraged to collect sufficient data to support development of
laboratory-specific criteria.
9.4.6	Corrective action for MS/MSD
If recovery or RPD limits are not met and the LCS, CCV and method blank are
within acceptable limits, this might be an indication of matrix interferences. If
sufficient sample is available, an MS/MSD should be reanalyzed, along with all
appropriate QC samples. If, after reanalysis, MS/MSD recovery limits are not
met, flag the results of the associated sample.
9.5 Laboratory Control Sample (LCS)
An LCS consists of an aliquot of clean reference matrix, of the same weight or volume as
the corresponding field samples and spiked with the same compounds at the same
concentrations used to spike the MS/MSD. The nominal volume of DCM added to water
samples is 2.0 mL. Because target compound and surrogate spiking solutions also
19
September 2016

-------
Analytical Protocol for VX Using GC/MS
contain DCM, the total volume of DCM added may be slightly greater than 2.0 mL. The
actual total volume of DCM must be used in the calculation in Section 12.2. When the
results of the MS/MSD analysis indicate matrix interference might be present, the LCS
results are used to verify that the interferences are due to the sample matrix and not from
artifacts introduced in the laboratory.
9.5.1	Preparation of LCS
Extract and analyze the LCS according to the procedures in Sections 11.2 (for
water samples), Sections 11.3 for (solid samples), or Section 11.4 (for wipe
samples).
9.5.2	Frequency of LCS analyses
One LCS should be prepared, extracted, analyzed, and reported for every 20 or
fewer field samples extracted in a batch of a similar matrix. The LCS must be
extracted and analyzed concurrently with the samples, using the same extraction
procedure, cleanup procedure (if required), and instrumentation.
9.5.3	Calculations for LCS
Calculate the recovery of each compound in the LCS using Eq. 11 (Section
12.2.8.2).
9.5.4	Technical acceptance criteria for LCS analysis
9.5.4.1	All LCSs should be extracted and analyzed at the frequency described
in Section 9.5.2 on a GC/MS system meeting the tuning, initial
calibration, CCV, and method blank technical acceptance criteria.
9.5.4.2	The limits for LCS compound recovery limits are 50 - 150 %.
9.5.5	Corrective action for LCS
9.5.5.1	If LCS recovery limits are not met, inspect the system for problems
and take corrective actions to achieve the acceptance criteria.
9.5.5.2	If LCS recovery limits cannot be met, flag all associated sample and
blank data accordingly.
9.6 Instrument Detection Limit (IDL) Determination
Laboratories may determine an IDL on each instrument used for analysis. While
determining IDLs is not required, IDL results can be helpful in determining an
appropriate spike level for use in determining the MDL (Section 9.7), as well as
instrument sensitivity to the target analyte. It is recommended that IDLs be verified
annually thereafter, or after major instrument maintenance. Major instrument
maintenance includes, but is not limited to: cleaning or replacement of the mass
spectrometer source, mass filters, or electron multiplier, or installing a different GC
column type. An IDL is instrument-specific and independent of sample matrices.
20
September 2016

-------
Analytical Protocol for VX Using GC/MS
9.6.1	An IDL is determined as the concentration that produces an average signal-to-
noise ratio (S:N) of between 3:1 and 5:1 for at least three replicate injections.
9.6.2	All documentation for the IDL determination shall be maintained at the
laboratory and provided to the reference agency or the data user upon request.
9.7	Method Detection Limit (MDL) Determination
Before any field samples are analyzed, a laboratory MDL should be determined in
appropriate reference matrices (i.e., reagent water, Ottawa sand, or clean wipes), using
the sample preparation and analytical procedures described in this protocol for each
specific matrix, and following the instructions and requirements described in 40 CFR Part
136, Appendix B.
9.7.1	The laboratory must use full method procedures to prepare and analyze at least
seven replicates.
9.7.2	Spike each replicate sample at concentrations of 1 - 5 times the IDL
concentration and analyze the samples following protocol procedures. The total
volume of DCM added to water samples will be slightly greater than the 2 mL
needed for extraction, and includes the volumes added for spiking VX and
surrogates.
9.7.3	To determine an MDL, the following equation is applied to the analytical results
(Student's t-factor depends on the number of replicates used; a factor of 3.14
assumes seven replicates):
Eq. 2 Method Detection Limit (MLD)
MDL = 3 .14 xsd
where:
sd = the standard deviation for the analytical results, and
3.14 = the Student's t-value for seven replicate samples
9.7.4	The MDL result calculated using the equation in Section 9.7.3 must meet the
following requirements as well as all other requirements specified in 40 CFR
Part 136, Appendix B:
•	MDL result must not be greater than the spiking level used for the MDL
determination.
•	MDL result must not be less than 0.10 times the spiking level used for the
MDL determination.
If either requirement is not met, the laboratory must adjust their spiking level
appropriately and repeat the MDL determination.
9.8	Quantitation Limit (QL) Determination
A QL determination is recommended for each laboratory/technician performing the
method for the first time, or in cases where new or repaired instrumentation is being used.
21
September 2016

-------
Analytical Protocol for VX Using GC/MS
Laboratory QLs are determined by first assessing at least four samples containing VX at
the level of the lowest calibration standard, against the criteria listed below. If any of
these criteria are not met, samples are assessed at concentrations of the next (second
lowest) calibration standard. These criteria are provided as guidance. If the criteria
cannot be met, the laboratory should consult analytical data requestor to determine if the
QL is sufficient to address project needs:
•	Results from spikes at the QL should be above the MDL.
•	The QL should be at or above the lowest calibration level.
•	The QL should be at least two times the MDL.
•	The RSD of results from spikes at the QL should be less than 2 0%.
•	The mean recovery of spikes at the QL should be within 50 - 150 %.
10.0 CALIBRATION AND STANDARDIZATION
10.1 Instrument Operating Conditions
10.1.1 GC
The following GC analytical conditions were used during laboratory studies and
are provided for guidance. Other conditions may be used, provided that all
technical acceptance criteria are met. Optimize GC conditions for analyte
separation and sensitivity. Once optimized, the same GC conditions must be
used for the analysis of all standards, samples, blanks, and MS/MSDs.
10.1.1.1 GC - Full-scan quadrupole
Initial column temperature:
Column temperature program:
Final column temperature hold:
Injector temperature:
Injection mode:
Sample injection volume:
GC column:
Column dimensions:
Carrier gas:
10.1.1.2 GC-TOF
Initial oven temperature:
Column temperature program:
Final column temperature hold:
40 °C for 3 minutes
40 - 150 °C at 10 °C/minute
150 - 280 °C at 25 °C/minute
280 °C; 10.8 minutes after the last
compound (triphenyl phosphate)
has eluted
250 °C
Grob-type, splitless for 0.75
minutes
1.0 jaL
Bonded phase silicon coated fused
silica capillary (see Section 6.4.2)
30 m x 0.25 mm (or 0.32 mm) x
0.25 (im
Helium at 32 cm/second
55 °C for 0.5 minutes
20 °C/minute to 100 °C (0
minute), 40 °C/minute to 280
280 °C (2.75 minutes)
22
September 2016

-------
Analytical Protocol for VX Using GC/MS
Inj ector temperature:
Injection mode:
Sample injection volume:
GC column:
Column dimensions:
Carrier gas:
250 °C
Grob-type, splitless
1.0 jiL
HP-5MS UI (Agilent) or
equivalent
15m X 0.18mm X 0.18(im
Helium at 1.2 mL/minute
10.1.2 MS
The following MS analytical conditions were used during laboratory studies and
are provided for guidance. Other conditions may be used, provided that all
technical acceptance criteria are met. Optimize MS conditions for analyte
separation and sensitivity. Once optimized, the same MS conditions must be
used for the analysis of all standards, samples, blanks, and MS/MSDs.
10.1.2.1	MS - Full-scan quadrupole
MS transfer line temperature:
Source temperature:
MS quadrupole temperature:
Electron energy:
Scan range:
Ionization mode:
Scan time:
Library searching:
10.1.2.2	MS-TOF
MS transfer line temperature:
Source temperature:
Electron energy:
Scan range:
Ionization mode:
Scan time:
280 °C
230 °C or according to
manufacturer's specifications
150 °C
70 eV (nominal)
35 to 500 m/z
Electron Ionization (EI), positive
3.15 scan/sec (minimum of 3 scans/
second)
NIST 05 Mass Spectral Data Base
295 °C
250 °C or according to
manufacturer's specifications
70 eV (nominal)
35 to 500 m/z
Electron Ionization (EI), positive
15 scans/second
10.2 GC/MS Mass Calibration (Tuning) and Ion Abundance
10.2.1 Summary of GC/MS instrument performance check
The GC/MS system must be tuned to meet the manufacturer's specifications,
using a suitable calibration compound such as perfluoro-tri-«-butylamine (FC-43)
or perfluorokerosene. The mass calibration and resolution of the GC/MS system
are verified by the analysis of the instrument performance check solution
(Section 7.2.2.3). Prior to the analysis of any samples, including MS/MSDs,
blanks, or calibration standards, the laboratory must establish that the GC/MS
system meets the mass spectral ion abundance criteria for the instrument
performance check solution (Table 3, Section 17) containing DFTPP.
23
June 2016

-------
Analytical Protocol for VX Using GC/MS
10.2.2 Frequency of GC/MS instrument performance check - The instrument
performance check solution must be injected once at the beginning of each 24-
hour period, during which samples, blanks, or standards are to be analyzed. The
24-hour period begins at the moment of injection of the DFTPP solution. The
time period ends after 24 hours have elapsed according to the system clock.
10.2.3 GC/MS instrument performance check
The analysis of the instrument performance check solution may be performed as
an injection of 50 ng or less of DFTPP into the GC/MS or by adding a sufficient
amount of DFTPP to the calibration standards to result in an on-column amount
of 50 ng or less of DFTPP (Section 7.2.2.3) and analyzing the calibration
standard.
10.2.4 Technical acceptance criteria for GC/MS instrument performance check
10.2.4.1	The instrument performance check solution must be analyzed at the
frequency described in Section 10.2.2.
10.2.4.2	Abundance criteria are listed in Table 3, Section 17 for guidance.
The mass spectrum of DFTPP must be acquired in the following
manner: three scans (the peak apex scan and the scans immediately
preceding and following the apex) are acquired and averaged.
Background subtraction is required, and must be accomplished using
a single scan acquired no more than 20 scans prior to the elution of
DFTPP. The background subtraction should be designed only to
eliminate column bleed or instrument background ions. Do not
subtract part of the DFTPP peak.
Note 1: All subsequent standards, samples, MS/MSDs, and blanks
associated with a DFTPP analysis must use the identical GC/MS
instrument run conditions.
Note 2: The above tuning criteria are suggested when using DFTPP.
If alternative tuning methods are used, consult the method or
manufacturer notes for guidance on criteria.
10.2.5 Corrective action for GC/MS instrument performance check
The following corrective actions are minimum procedures. The analyst may try
other corrective action procedures to meet criteria.
10.2.5.1	If the GC/MS instrument performance check technical acceptance
criteria are not met, re-tune the GC/MS system. It may be necessary
to perform maintenance to achieve the technical acceptance criteria.
10.2.5.2	The instrument performance check technical acceptance criteria in
Section 10.2.4 must be met before any standards, samples, including
MS/MSDs, or required blanks are analyzed.
24
September 2016

-------
Analytical Protocol for VX Using GC/MS
10.3 Initial Calibration
Prior to sample analysis and after instrument performance check technical acceptance
criteria have been met, each GC/MS system must be calibrated at a minimum of five
concentrations (Section 7.2.2.4.1 and Table 8, Section 17) to determine instrument
sensitivity and the linearity of GC/MS response for the target and surrogate compounds.
If the RSD criteria cannot be met, a linear or quadratic curve may be used. Each initial
calibration standard contains the target compound, surrogate, and internal standard.
10.3.1	Frequency of initial calibration
10.3.1.1	Each GC/MS should be calibrated whenever the laboratory takes
corrective action that might change or affect the initial calibration
criteria, or if the CCV technical acceptance criteria are not met.
10.3.1.2	If time remains in the 24-hour period after meeting initial calibration
acceptance criteria, samples may be analyzed. In this case, it is not
necessary to analyze an opening CCV standard prior to sample
analysis.
10.3.2	Procedure for initial calibration
10.3.2.1	Prepare at least five calibration standards containing the detected
target compound and associated surrogate. Example concentrations
for the calibration standards are provided in Section 7.2.2.4.1 and
Table 8, Section 17.
10.3.2.2	Add a sufficient amount of internal standard solution (Section 7.2.2.5)
to aliquots of calibration standards to result in 0.5 ng/(.iL of the
internal standard. Standards specified in Section 7.2.2.4 should
permit the target compound to have a relative retention time (RRT) of
approximately 0.60 to 1.70, using the internal standard listed in Table
4, Section 17.
10.3.2.3	Analyze each calibration standard by injecting 1.0 |_iL of standard.
The same injection volume must be used for all standards, samples,
and blanks.
10.3.3	Calculations for initial calibration
10.3.3.1 Calculate the relative response factors (RRFs) for VX and the
surrogate using Eq. 3 and the primary characteristic ions found in
Table 5, Section 17. For the internal standard, use the primary ion
listed in Table 5 unless interferences are present (e.g., peak overlap,
co-elution). Unless otherwise stated, the area response of the primary
characteristic ion is the quantitation ion.
Eq. 3 RRF Calculation
A C
RRF = —— x ——
A is Cx
25
September 2016

-------
Analytical Protocol for VX Using GC/MS
where:
Ax = Area of the characteristic ion for the compound to be measured
(Table 5, Section 17)
Ais = Area of the characteristic ion for specific internal standard
(Table 5)
Cis = Amount of the internal standard injected (ng)
Cx = Amount of the target compound or surrogate injected (ng)
10.3.3.2 The Mean RRF ( RRF) for the Initial Calibration
RRFs and mean RRFs must be calculated for all compounds.
Calculate the %RSD of the RRF values for the initial calibration.
10.3.4 Technical acceptance criteria for initial calibration
10.3.4.1	An initial calibration should be performed at the frequency described
in Section 10.3.1 on a GC/MS system meeting the instrument
performance check technical acceptance criteria (Section 10.2.4).
10.3.4.2	The RRFs for VX and the surrogate should be greater than or equal to
0.01.
10.3.4.3 The %RSD of the RRFs over the initial calibration range for VX and
each surrogate must be less than or equal to 20. If %RSD cannot
meet this criterion, curve fitting by linear or quadratic regression may
be used provided the R2 value is greater than or equal to 0.99 (linear)
or 0.995 (quadratic). Refer to Section 12.2.7 if linear regression is
used; refer to SW-846 Method 8000C (Reference 16.16), if quadratic
curve fitting is needed. If regression curve fitting is used, percent
drift (PD) (as calculated using Eq. 4, Section 10.4.4.2) for each
standard must be less than or equal to ±50.
10.3.5 Corrective action for initial calibration
The following corrective actions are minimum procedures. The analyst may try
other corrective action procedures to meet criteria.
10.3.5.1 If technical acceptance criteria using at least one of the three optional
approaches to initial calibration (%RSD of the RRFs, linear
regression, or quadratic regression) are not met, inspect the system for
problems, take corrective actions, remake standards and re-calibrate.
If criteria are not met with re-calibration, remake the calibration
standards and repeat. If the criteria are still not met, the laboratory
will flag all data associated with the calibration.
Note: If technical acceptance criteria for the initial calibration are
not met and the initial calibration contains more than five calibration
levels, the laboratory may remove calibration point(s) from either
extreme end of the calibration range and reassess the calibration.
Data points from within a calibration range must not be removed.
10.3.5.2 Initial calibration technical acceptance criteria must be met before any
26	September 2016

-------
Analytical Protocol for VX Using GC/MS
samples, including MS/MSDs or required blanks are analyzed and
reported without data qualification.
10.4 Continuing Calibration Verification (CCV)
10.4.1 Summary of CCV
Prior to the analysis of samples and after instrument performance check and
initial calibration technical acceptance criteria have been met, each GC/MS
system must be routinely checked by analyzing a CCV standard to ensure that the
instrument continues to meet the instrument sensitivity and linearity
requirements. The CCV standard contains VX, surrogate, and internal standard.
10.4.2	Frequency of CCV - Each GC/MS used for analysis must be checked at the
beginning and at the end of each analytical batch of <20 injections, excluding
instrument blanks. When subsequent analytical batches are run within a single
24-hour period, the closing CCV may be used as the opening CCV for a new
analytical batch, provided that the closing CCV meets all technical acceptance
criteria for an opening CCV (see Section 10.4.5).
10.4.3	Procedure for CCV
10.4.3.1	Add a sufficient amount of internal standard solution (Section 7.2.2.5)
to an aliquot of CCV standard to result in a concentration of 0.5
ng/(.iL for both quadrupole and TOF analyses.
10.4.3.2	Analyze the CCV standard by injecting 1.0 |_iL of standard.
10.4.4 Calculations for CCV
Calibration verification involves calculation of the PD (Eq. 4a) or the percent
difference of the RRFs between the initial calibration and each subsequent CCV
(Eq. 4b). The CCV approach will depend on the how the initial calibration was
performed. If a regression technique (linear or quadratic) was used, then Eq. 4a
is used to determine a percent drift. If the RRF approach is used, then Eq. 4b is
used to determine the percent difference of the RRFs.
10.4.4.1 If regression techniques are used for initial calibration, the CCV must
be evaluated in terms of PD, which is calculated using concentrations
(see Eq. 3a).
Eq. 4a Percent Drift (PD) Calculation for CCV
Calculated Concentration-Theoretical Concentration
lJL) =	xl00%
Theoretical Concentration
10.4.4.2 Calculate an RRF for the target compound and surrogate using Eq. 3
and the primary quantitation ions found in Table 5, Section 17. If
regression techniques are used for the initial calibration, proceed to
Section 10.4.4.3.
27
September 2016

-------
Analytical Protocol for VX Using GC/MS
10.4.4.3 Calculate the Percent Difference (%Difference) between the RRF of
the most recent initial calibration and the continuing calibration
verification RRF for each target compound and surrogate using Eq.
4b.
Eq. 4b Relative Response Factor (RRF) % Difference
Calculation
RRF — RRF
%DifferenceRRF =	0 —L x 100
RRF RRF,
where:
RRFi= Mean RRF from the most recent initial calibration meeting
technical acceptance criteria.
RRFC = RRF from CCV standard.
10.4.5 Technical acceptance criteria for CCV
10.4.5.1	The CCV standard should be analyzed at or near the mid-point
concentration level, at the frequency described in Section 10.4.2, on a
GC/MS system meeting the instrument performance check and the
initial calibration technical acceptance criteria.
10.4.5.2	The RRF for VX and the surrogate must be greater than or equal to
0.01.
10.4.5.3	For the opening CCV, the PD or %Difference of RRFs for the target
compound and surrogate should be within the range of ±40.
10.4.5.4	For the closing CCV, the PD or %Difference of RRFs for the target
compound and surrogate should be within the range of ±50.
10.4.5.5	Excluding those ions in the solvent front, no quantitation ion may
saturate the detector.
10.4.6 Corrective action for CCV
The following corrective actions are minimum procedures. The analyst may try
other corrective action procedures to meet criteria.
10.4.6.1	If the CCV technical acceptance criteria in Section 10.4.5 are not met,
recalibrate the GC/MS instrument according to Section 10.3.
10.4.6.2	CCV technical acceptance criteria should be met before any samples
MS/MSDs, or required blanks are analyzed. If CCV criteria are not
met, flag associated samples and blanks accordingly.
28
September2016

-------
Analytical Protocol for VX Using GC/MS
10.5 Instrument Blank
10.5.1	Summary of instrument blank
An instrument blank is comprised of DCM spiked with internal standard at the
same concentration used for associated samples. The purpose of the instrument
blank is to investigate the impact of carry-over.
10.5.2	Frequency of instrument blank
An instrument blank is recommended for analysis following suspected carry-over
or during analysis of samples containing suspected high concentrations.
10.5.3	Procedure for instrument blank analysis
Add sufficient amount of internal standard solution (Section 7.2.2.5) to an aliquot
of the solvent used to prepare calibration standards and sample extracts to result
in a concentration of 0.5 ng/|_iL. Analyze each instrument blank by injecting a
volume of 1.0 (iL.
10.5.4	Calculations for instrument blank
Calculate the concentrations of any observed target analyte using Eq. 6 (Section
12.2.6.1), setting Vt, V0, and dilution factor (DF) all equal to 1.
10.5.5	Technical acceptance criteria for instrument blank
If an instrument blank is analyzed, the concentration of VX in the instrument
blank should be less than the concentration of VX in the low calibration standard.
The area response of the internal standard should be within 50 - 150 % of the
associated CCV or mid-level concentration of the initial calibration.
10.5.6	Corrective action for instrument blank
If an instrument blank is analyzed and the instrument blank technical acceptance
criteria are not met, analyze an additional instrument blank. If the problem
persists, inspect the system for problems and take corrective actions to achieve
the acceptance criteria. Instrument blank technical acceptance criteria should be
met before samples are analyzed. Samples that are analyzed with corresponding
instrument blanks that do not meet the instrument blank criteria should be re-
analyzed, or the corresponding data should be flagged.
29
September 2016

-------
Analytical Protocol for VX Using GC/MS
11.0 ANALYTICAL PROCEDURE
11.1	Sample Preparation - General
11.1.1	If less than the specified sample amount is received, the laboratory should use a
reduced sample size for the analysis and adjust the calculation accordingly. For
the purposes of this method, it is recommended that some sample be retained (if
possible) for potential future evidentiary use.
11.1.2	If multi-phase samples (e.g., a two-phase liquid sample, oily, sludge/sandy soil
sample) are received by the laboratory, the laboratory shall contact the agency to
apprise them of the type of sample received. If some or all phases of the sample
are amenable to analysis, the agency requesting the analyses may require the
laboratory to do any of the following:
•	Mix the sample and analyze an aliquot from the homogenized sample.
•	Separate the phases of the sample and analyze each phase separately.
•	Separate the phases and analyze one or more of the phases, but not all of the
phases.
•	Do not analyze the sample.
11.2	Preparation of Water Samples Using Microscale Extraction
11.2.1 Approximately 35 mL of a water sample is required for this extraction. If
extraction is to be performed in the sample receipt vial, remove any excess
sample such that a total sample volume of 35 mL is obtained and recap the vial.
Weigh the capped vial. Record the weight to the nearest 0.1 gram.
Alternatively, 35.0 mL of sample can be transferred by pipette into the vial and
the weighing step can be eliminated.
Note: The conical bottoms of centrifuge vials allow the DCM layer to be
removed more easily than the VOA vials.
11.2.2	pH determination - If pH determination is requested, transfer a separate aliquot
of the water sample to a beaker, and determine the pH of the sample with a pH
meter or wide-range pH paper. Document the pH in the data narrative. Discard
the sample aliquot.
11.2.3	For GC/MS full-scan and TOF analysis, spike l .0 |ig of each surrogate (add 40
(.iL if prepared as in Section 7.2.2.1) into each sample, blank, etc. The final
volume of the extracts will be 2.0 mL; therefore, the concentration of the
surrogates in the extract is expected to be 0.50 |_ig/mL.
11.2.4	Add -8.8 grams of sodium chloride and shake vigorously, or vortex each vial for
two minutes or until the sodium chloride dissolves completely.
11.2.5	Add 2.00 mL of DCM, using a Class A volumetric pipette or syringe. Cap
tightly and agitate the contents vigorously for approximately two minutes, either
by hand or using a vortex mixer or shaker table.
30
September 2016

-------
Analytical Protocol for VX Using GC/MS
Briefly allow the phases to settle by gravity for ~ five minutes. Centrifugation
(three minutes at 500 revolutions per minute [rpm]) is strongly recommended to
facilitate separation of the phases and affords a greater recovery of the final
sample extract. CAUTION: The maximum safe handling speed of each
centrifuge will depend, in part, on the vials used and should be determined prior
to use.
Using a 2.0-mL syringe or pipette, transfer approximately 1.0 mL (or as much as
possible) of the DCM (lower) layer to a 2-mL or 4-mL vial with a PTFE-lined
screw cap, taking precautions to exclude any water from the syringe or pipette.
Add a small amount (-50 mg) of anhydrous sodium sulfate to the vial, then cap
the vial, and shake vigorously or vortex for two minutes. Make sure that the
extract is sufficiently dry and that some of the sodium sulfate added is free-
flowing (i.e., not clumped).
Using a 1.0 mL syringe or pipette, transfer 1.0 mL (or a known volume, if less
than 1.0 mL of extract is collected) of the dried extract to a 2.0-mL vial (or
autosampler vial insert) with a PTFE-lined screw cap. Cap the vial.
Note: If stored prior to analysis, extracts must be protected from light and stored
at<6°C (Section 8.3).
11.2.9	Discard the remaining contents of the VOA vials according to laboratory waste
disposal guidelines. Shake off the last few drops with short, brisk wrist
movements. If needed, rinse the vial with a water-soluble solvent to ensure that
the extraction solvent is removed. If the vial was pre-weighed (i.e., exact
sample volume used in Section 11.2.1 is unknown), reweigh the capped vial, and
record the weight to the nearest 0.1 grams. The difference between this weight,
and the weight determined in Section 11.2.1 is equal to the volume of water
extracted, in milliliters. As the density of water is 1.00 g/mL (at 20°C), the
volume of water extracted, in milliliters, may be assumed to be equal to the
weight of water extracted.
11.2.10	Proceed to Section 11.6 for sample analysis.
11.3 Preparation of Solid Samples Using Microscale Extraction
Note: The following procedures have been evaluated in a multi-laboratory study using
Ottawa sand and dried soils, and have not been evaluated for field samples. Laboratory
results are provided in Section 17.
11.3.1	Decant and discard any water layer. Mix samples thoroughly, especially
composited samples. Discard any foreign objects such as sticks, leaves, and
rocks.
11.3.2	pH determination - If pH determination is requested, transfer a 1:1 (w:w) ratio
of sample:water to a 100-mL beaker and stir for one hour. Determine the pH of
the sample with a pH meter or wide-range pH paper, and document the pH in the
data narrative. Discard this portion of the sample.
11.2.6
11.2.7
11.2.8
31
September 2016

-------
Analytical Protocol for VX Using GC/MS
11.3.3 Percent moisture determination - If percent moisture determination is requested,
immediately after weighing the sample for extraction, weigh 5-10 grams of the
sample into a tared vial. Dry overnight at 103 - 105 °C, and cool in a desiccator
before weighing. Determine the percent moisture (%Moisture) using Eq. 5.
CAUTION: Due to the high toxicity associated with VX, percent moisture
determinations should be performed only in an oven with appropriate engineering
controls.
Eq. 5 Percent Moisture Calculation
vet sample
grams of wet sample
%Moistu,e = gramS °f WCt Samp'e " gnmS °f iry' Smlp'e x 100
11.3.4 Extraction of VX from soil samples
11.3.4.1 Weigh 10 grams of sample into a tared extraction vial (i.e., 60-mL
VOA vial). Wipe the lip and threads of the vial with a clean cloth
(e.g., Kimwipe®[Kimberly-Clark Professional, Roswell, GA] or
equivalent). Record weight to the nearest 0.01 gram.
11.3.4.2	For full-scan quadrupole and TOF MS, add 0.5 |ig surrogate (add 20
(.iL if using a 25 |ig/m L surrogate spiking solution) in DCM to the
vial. The final volume of the extract is 1.0 mL; therefore, the
concentration of surrogates in the extract is expected to be 0.5 |_ig/mL.
11.3.4.3	Add 5-10 pre-cleaned glass beads. Mix until the sample is
homogenized using a metal spatula. Break up any chunks with a
metal spatula, working quickly but gently.
11.3.4.4	Add 30 mL of Tris buffer (Section 7.1.4) to the vial, and cap tightly.
11.3.4.5	Shake the vial vigorously or vortex for approximately 20 seconds or
until the slurry is free-flowing. Break up any chunks with a metal
spatula, working quickly but gently. Cap immediately when finished.
11.3.4.6	Extract the sample by agitating for approximately 15 minutes, using a
shaker table or sonicator.
Note: Sonication at high power should be avoided for soils having
high silt content; the resulting fine particles can create problems with
filtering and extracting solvent from the sample.
11.3.4.7 Allow the solids to settle or centrifuge for 1 - 2 minutes at 1000 rpm.
If the solid is still unsettled, repeat the centrifuge step, but increase
speed to 2500 rpm. CAUTION: The maximum safe handling speed
of each centrifuge will depend, in part, on the vials used and should
be determined prior to use. Repeat until the solid is completely
settled. If after repeating the centrifuging steps several times the solid
is still unsettled, proceed to Section 11.3.4.8. Once the solid has
32
September 2016

-------
Analytical Protocol for VX Using GC/MS
settled, decant or pipette the solvent layer into a pre-cleaned, 60-mL
VOA vial with PTFE-lined screw cap and proceed to 11.3.4.9.
Note: For solids that have difficulty settling, pipetting is
recommended.
11.3.4.8	If solids are not settled out by centrifugation (Section 11.3.4.7), filter
by placing a small plug of glass wool into a small glass funnel. Wet
the glass wool thoroughly with Tris buffer. Pour the sample (both
solid and liquid portions) into the funnel. Rinse with approximately 2
- 3 mLs of Tris buffer as soon as the surface is exposed, not allowing
it to dry.
11.3.4.9	Add 5.00 mL of DCM, using a Class A volumetric pipette or syringe.
Cap tightly, and agitate by vigorous shaking, vortex mixer, shaker
table, or sonication for approximately two minutes.
11.3.4.10	Briefly allow the phases to settle by gravity or by centrifugation for
~5 minutes.
Note: Centrifugation (three minutes at 500 rpm) is strongly
recommended to facilitate separation of the phases and affords a
greater recovery of the final sample extract.
11.3.4.11	Using a syringe or pipette, transfer as much of the DCM (lower) layer
as possible to a vial with a PTFE-lined screw cap, taking precautions
to exclude any water from the syringe or pipette. Add a small amount
(-50 mg) of anhydrous sodium sulfate to the vial, cap the vial, and
shake vigorously or vortex for two minutes. Make sure that the
extract is sufficiently dry and that some of the sodium sulfate added is
free-flowing (e.g., not clumped).
11.3.4.12	Proceed to Section 11.5 for extract concentration. Once the extract is
concentrated, proceed to Section 11.6 for analysis.
11.4 Preparation of Wipe Samples by Microscale Extraction
11.4.1	Place the wipe into an extraction vial (i.e., 40-mL VOA vial). For GC/MS full-
scan and TOF analyses, add 0.5 |ig of each surrogate standard compound (add 20
(.iL if prepared as in Section 7.2.2.1) in DCM directly onto the wipe. The final
volume of the extract is 1.0 mL; therefore, the concentration of the surrogates in
the extract is expected to be 0.5 |_ig/mL.
11.4.2	Add 15 mL of DCM to the vial and cap tightly.
11.4.3 Extract the sample by agitating for approximately 15 minutes, using a shaker
table or sonicator.
11.4.4 Remove the vials from the sonicator or shaker table, shake briefly by hand and
allow the solvent layer to settle. Transfer the solvent layer by pipette into a pre-
cleaned, 40-mL, clear glass vial with PTFE-lined screw cap.
33
September 2016

-------
Analytical Protocol for VX Using GC/MS
11.4.5 Proceed to Section 11.5 for extract concentration. Once the extract is
concentrated, proceed to Section 11.6 for analysis.
11.5 Final Concentration of Extract - Nitrogen Evaporation Technique (RapidVap® or
equivalent) for solid and wipe samples.
11.5.1 Nitrogen evaporation technique - If using a RapidVap® or TurboVap® N2
systems (Biotage, LLC, Charlotte, NC), follow the manufacturer's guidelines.
For the RapidVap®, a temperature of 40 °C is recommended. If using a water
bath, place the vial in a warm water bath (30-40 °C recommended) and
evaporate the solvent volume to just below 1.0 mL by blowing a gentle stream of
clean, dry nitrogen above the extract. It is recommended that the internal wall of
the vial be rinsed down several times with DCM during the operation. If using a
RapidVap® the solvent rinse should, at a minimum, be done during final
adjustment of the extract volume. During evaporation, the tube solvent level
must be kept below the water level of the bath. The extract must never be
allowed to become dry. Adjust the final volume to 1.0 mL with the same solvent
used for extraction. Transfer the extract to a 2-mL autosampler vial, cap, and
label the vial. Store at 4 °C (± 2 °C).
CAUTION: Gas lines from the gas source to the evaporation apparatus should
be stainless steel, copper, or PTFE tubing. With the exception of PTFE, plastic
tubing must not be used between the carbon trap and the sample since it can
introduce interferences.
11.5.2 Final extract volumes - The final extract volumes in Sections 11.5.2.1 through
11.5.2.4 are recommended volumes.
11.5.2.1 Water/Liquid - As concentration of the sample extract is not needed
for these sample matrices, no adjustment of the final extract volume is
required. The nominal volume of DCM added to water samples is 2.0
mL. The target compound and surrogate spiking solutions also
contain DCM; therefore, the total volume of DCM added may be
slightly greater than 2.0 mL. The actual total volume of DCM should
be used in the calculations in Section 12.2.
11.5.2.2 Solid - Adjust the extract volume to a final volume of 1.0 mL with
DCM.
11.5.2.3	Wipe - Adjust the extract volume to a final volume of 1.0 mL with
DCM.
11.5.2.4	If extracts are stored prior to analysis, transfer the extract to a PTFE-
lined screw-cap vial (approximately 2.0 mL), label the vial, and store
at <6 °C.
11.6 Extract Analysis by GC/MS
11.6.1 Analyze extracts only after the GC/MS system has met the instrument
34
September 2016

-------
Analytical Protocol for VX Using GC/MS
performance check (Section 10.2.4), initial calibration (Section 10.3.5), and CCV
technical acceptance requirements (10.4.5). The same instrument conditions
used for calibration must be used for the analysis of samples.
11.6.2	Add internal standard solution (Section 7.2.2.5) to an accurately measured
aliquot of each extract, cap the vial, and invert several times to mix the contents.
For full-scan quadrupole or TOF MS analyses, add a sufficient amount of
internal standard solution to result in 0.5 ng/(.iL concentration of the internal
standard.
11.6.3	If extracts are to be diluted, add the internal standard after dilution. Internal
standard must be added to maintain the required 0.5 ng/(.iL (for both full-scan
quadrupole and TOF) of internal standard in the extract volume.
11.6.4	Inject 1.0 |_iL of the extract into the GC/MS.
Note: The same injection volume used for calibration standards must be used
for extracts.
11.6.5 Sample extract dilution
11.6.5.1	If the response of the target compound in any extract exceeds the
response of the target compound in the high standard of the initial
calibration, that extract must be diluted. Add additional internal
standard solution such that the concentration in the diluted extract is
0.5 ng/(.iL for each internal standard, and analyze the diluted extract.
11.6.5.2	Use the results of the original analysis to determine the approximate
dilution factor (DF) required for the analyte peak to fall within the
initial calibration range. The DF chosen must keep the response of
the peak for the target compound in the upper half of the calibration
range of the instrument.
12.0 CALCULATIONS AND DATA ANALYSIS
12.1 Qualitative Identification of Target Compounds
12.1.1	The target compound should be identified by an analyst competent in the
interpretation of mass spectra by comparison of the sample mass spectrum to the
mass spectrum of the standard of the suspected compound. Two criteria must be
satisfied to verify identification:
1)	Elution of the sample analyte within the GC RRT unit window
established from the 24-hour calibration standard; and
2)	Correspondence of the sample analyte and calibration standard
component mass spectra.
12.1.2	For establishing correspondence of the GC RRT, the sample component must
compare within ±0.06 RRT units of the standard component. For samples
analyzed during the same 24 hour time period as the initial calibration standards,
35
September 2016

-------
Analytical Protocol for VX Using GC/MS
compare the analyte RTs to those from the midpoint initial calibration standard.
Otherwise, use the corresponding CCV standard. If coelution of interfering
components prohibits accurate assignment of the sample component RRT from
the total ion chromatogram, the RRT should be assigned by using EICPs for ions
unique to the component of interest (see Table 5, Section 17 for appropriate
characteristic ions, surrogate, and internal standard).
12.1.3	For comparison of standard and sample component mass spectra, mass spectra
obtained from a calibration standard at a concentration of the target analyte
closest to the concentration of the analyte in the sample are required. Once
obtained, these standard spectra may be used for identification purposes only if
the GC/MS meets the DFTPP instrument performance requirements (see Section
10.2 for instrument performance check requirements).
12.1.4	For TOF MS and full-scan quadrupole MS analyses, all ions present in the
standard mass spectrum at a relative intensity greater than 10 % (the most
abundant ion in the spectrum equaling 100 %) must be present in the sample
spectrum. The relative intensities of ions specified in Table 5, Section 17 must
agree within ± 20 % between the standard and sample spectra (e.g., for an ion
with an abundance of 50 % in the standard spectra, the corresponding sample ion
abundance must be between 30 - 70 %). Ions greater than 10 % in the sample
spectrum, but not present in the standard spectrum, must be considered and
accounted for by the analyst making the comparison. The verification process
should favor false positives. All compounds meeting the identification criteria
must be reported with their spectra. When target compounds are below QLs but
the spectrum meets the identification criteria, report the concentration with a "J".
For example, if the QL is 5.0 (ig/L and concentration of 3.0 |ig/L is calculated,
report as "3.0 J".
For TOF analysis, the signals for the quantitation ions in Table 5, Section 17
must be present and must maximize within a period of two seconds. The S:N for
the GC peak at each ion must be greater than or equal to 2.5 for the target
compound and surrogate detected in a sample extract, and greater than or equal to
10 for the target compound and surrogate in the CCV standard.
12.1.5	If the compound cannot be verified by all of the spectral identification criteria in
Sections 12.1.1-12.1.4, but in the technical judgment of the mass spectral
interpretation specialist the identification is correct, then the laboratory should
report the identification and proceed with quantitation.
12.2 Data Analysis and Calculations of Target Compounds
12.2.1	The target compound is quantitated by the internal standard method, using the
EICP area of primary quantification ions listed in Table 5, Section 17.
12.2.2	It is expected that situations will arise when the automated quantitation
procedures in the GC/MS software provide inappropriate quantitations. This
normally occurs when there is compound coelution, baseline noise, or matrix
interferences. In these circumstances, the laboratory must perform a manual
quantitation. Manual integrations are performed by integrating the area of the
quantitation ion of the compound. This integration should include the area
36
September 2016

-------
Analytical Protocol for VX Using GC/MS
attributable to the specific target compound. The area integrated must not
include baseline background noise. The area integrated should not extend past
the point where the sides of the peak intersect with the baseline noise. Manual
integration is not to be used solely to meet QC criteria, nor is it to be used as a
substitute for corrective action on the chromatographic system.
12.2.3	In some instances, the data system report may have been edited or manual
integration or quantitation may have been performed. In all such instances, the
GC/MS operator should identify such edits or manual procedures by initialing
and dating the changes made to the report, and include the integration scan range.
The GC/MS operator should also mark each integrated area on the quantitation
report.
12.2.4	The requirements listed in Sections 12.2.1 - 12.2.3 apply to all standards,
samples, and blanks.
12.2.5 The RRF from the initial calibration is used to calculate the concentration in the
sample. Secondary ion quantitation is allowed only when there are sample
interferences with the primary ion. If linear regression is used, a regression curve
should be used to calculate the concentration in samples. Refer to Section 12.2.7
for calculating sample concentration using linear regression.
12.2.6 Calculate the concentration in the sample using the RRF and Eq. 6-8.
12.2.6.1 Water
Eq. 6 Concentration of Water Sample
Concentration (|ig/L) = (A ) (I ) (V,) (DF)
(Ais) (RRF) (V0) (Vi)
where:
Ax = Area of the characteristic ion for the target compound
A1S = Area of the characteristic ion for the internal standard
Is = Amount of internal standard injected in ng
V0 = Volume of water extracted in mL
V, = Volume of extract injected in |_iL
Vt = Volume of the extract in |aL
(Extraction of water samples does not include
concentration; Vtis equal to the sum of the volumes of
solvent added for extraction and the addition of
surrogates and any spiked target compound.)
RRF = Mean RRF determined from the initial calibration
standard
DF = Dilution Factor. If no dilution is performed, DF = 1.0.
The DF for analysis of water samples is defined as:
(jL most conc. extract used to make dilution + jxL clean solvent
Dr —
(jL most conc. extract used to make dilution
37
September 2016

-------
Analytical Protocol for VX Using GC/MS
12.2.6.2	Solid
Eq. 7 Concentration of Solid Sample
Eq. 7 includes a %moisture factor (D) for those cases when data is to
be reported on the basis of dry sample weight. In cases where results
are reported in terms of sample weight, this factor is deleted from the
equation.
(A )(I )(V )(DF)
Concentration ug/Kg (Dry weight basis) =	=====	
(Ais)(Vi)(RRF)(Ws)(D)
where:
Ax, Is, Ais, V, and RRF are as given for water, above.
V, = Volume of concentrated extract in |.iL
^ 100 - %Moisture
100
%Moisture is as given in EQ. 5
Ws = Weight of sample extracted in grams
RRF = Mean RRF determined from the initial calibration
standard
DF = Dilution Factor
12.2.6.3	Wipes
Eq. 8 Concentration of Wipe Sample
(Ax)(ls)(V<)(PF)
Concentration „gW -	,a) ( RgJ )
where:
Ax = area response for the compound to be measured, counts
A1S = area response for the internal standard, counts
Is = amount of internal standard, |.ig
RRF = the mean RRF from the most recent initial calibration,
dimensionless
Area = area of surface wiped, cm2. If concentration is reported
as ng/wipe, area = 1 wipe.
V, = volume of concentrated extract, |.iL
V, = volume of extract injected, |.iL
DF = dilution factor for the extract. If there was no dilution, DF
equals 1. If the sample was diluted, DF is greater than 1.
38
September 2016

-------
Analytical Protocol for VX Using GC/MS
12.2.7	Calculate the concentration in the sample using linear regression.
The following procedure is used to calculate analyte concentrations using linear
regression calibration curves. Refer to SW-846 Method 8000C (Reference
16.16) if calibration curves were determined using quadratic equations.
12.2.7.1	Set y = (Peak Area of Target/Peak Area of Internal Standard) and x
= (Theoretical Concentration of Target/Theoretical Concentration of
Internal Standard).
12.2.7.2	Plot (Peak Area of Target/Peak Area of Internal Standard [Y-axis])
vs. (Theoretical Concentration of Target/Theoretical Concentration
of Internal Standard).
12.2.7.3	Determine the slope of the line (m) and the y-intercept (b).
12.2.7.4	Rearrange the line equation to solve for x: x = (y-b)/m.
12.2.7.5	Multiply x by the concentration of the internal standard to get
concentration of target analyte in extract.
12.2.7.6	Multiply the concentration of target analyte in the extract by the
extract volume, and divide by the sample volume to get
concentration of target analyte in sample.
12.2.8	Adjusted QL calculations
Adjusted QLs are used in situations when the laboratory may not have a sample
size that is sufficient for the method as written, or if the prescribed extract
volume was not used or recovered.
12.2.8.1 Water samples
Eq. 9 Aqueous Adjusted QL
Adjusted QL = Method QL x
(Vx)(Vt)(DF)
(Vo)(vc)
where:
Vt, DF, and V0 are as given in Eq. 6.
Vx = Recommended sample volume (35 mL)
Vc = Recommended concentrated extract volume (2000 jxL)
12.2.8.2 Solid samples
Eq. 10 Solid Adjusted QL
Adjusted QL = Method QL x
(Wx)(Vt)(DF)
(WS)(VC)(D)
where:
Vt and DF are as given in Eq. 6.
Ws and D are as given in Eq. 7.
39
September 2016

-------
Analytical Protocol for VX Using GC/MS
Wx = Method sample weight (10 grams)
Vc = Method concentrated extract volume (1000 |iL)
12.2.9 Surrogate recoveries
12.2.9.1	Calculate surrogate recoveries for all samples, blanks, and
MS/MSDs. Determine if recovery is within limits (Table 7,
Section 17).
12.2.9.2	Calculate the concentrations of the surrogates using the same
equations used for VX. Calculate the recovery of each surrogate
using Eq. 11.
Eq. 11 Percent Recovery
Cs
Recovery = %R = — x 100
where:
Cs = Measured concentration of the spiked sample aliquot.
Cn = Nominal (theoretical) concentration increase that results
from spiking the sample, or the nominal concentration of the
spiked aliquot (for LCS).
12.3 Technical Acceptance Criteria for Sample Analysis
12.3.1	The samples must be analyzed on a GC/MS system meeting the instrument
performance check, initial calibration, CCV, and blank technical acceptance
criteria.
12.3.2	The sample must be extracted and analyzed within the technical holding times.
12.3.3	The sample must have an associated method blank meeting the blank technical
acceptance criteria.
12.3.4	The percent recoveries of the surrogates in a sample should be within the
recovery limits listed in Table 7, Section 17 (see Table 4, Section 17 for analyte
specific surrogates). The surrogate recovery requirements do not apply to
samples that have been diluted.
12.3.5	The instrumental response (EICP area) for the internal standard in the sample
must be within the range of 50.0 - 200 % of the response of the internal standard
in the most recent CCV standard analysis.
12.3.6	The RT shift for the internal standard must be within ± 0.50 minute (30 seconds)
between the sample and the most recent CCV standard analysis.
12.3.7	Excluding those ions in the solvent front, no ion may saturate the detector. If a
target compound concentration exceeds the upper limit of the initial calibration
range, a more dilute aliquot of the sample extract must also be analyzed.
40
September 2016

-------
Analytical Protocol for VX Using GC/MS
12.4 Corrective Action for Sample Analysis
12.4.1	The sample technical acceptance criteria must be met before data are reported. If
the corrective actions described in this section did not solve the problem, all
associated sample and blank data must be flagged accordingly.
12.4.2	Corrective action for failure to meet instrument performance checks and initial
and continuing calibration verification must be completed before the analysis of
samples. If the corrective actions described in Sections 10.2.5 (for instrument
performance check), 10.3.5 (for initial calibration), or 10.4.6 (for CCV) did not
solve the problem, all associated sample and blank data must be flagged
accordingly.
12.4.3	Corrective Action for Surrogate Recoveries that Fail to Meet Their Acceptance
Criteria (Section 12.3.4 and Table 7, Section 17).
12.4.3.1	If the surrogate recoveries in a sample fail to meet the acceptance
criteria specified in Section 12.3.4, then check calculations, sample
preparation logs, surrogate standard spiking solutions, and the
instrument operation.
12.4.3.2	If the above actions do not correct the problem, then the problem
might be due to a sample matrix effect. To determine if there was
matrix effect, take the following corrective action steps:
12.4.3.2.1 Re-extract (if possible) and reanalyze the sample.
Note: Samples with corresponding MS and MSDs
should be re-extracted and reanalyzed only if surrogate
recoveries in the sample were considered unacceptable,
and the surrogate recoveries met the acceptance criteria
in both the corresponding MS and MSD.
12.4.3.2.2	If the surrogate recoveries meet acceptance criteria in the
re-extracted/re-analyzed sample, then the problem was
within the laboratory's control.
12.4.3.2.3	If surrogate recoveries are outside the acceptance criteria
in the reanalysis, flag the sample data for the associated
target compounds and submit data from both analyses.
Distinguish between the initial analysis and the
reanalysis on all data.
12.4.4 Corrective action for internal standard compound responses and/or RTs that fail
to meet the acceptance criteria (Sections 12.3.5 and 12.3.6)
12.4.4.1 If the internal standard in a sample fails to meet the acceptance
criteria, check calculations, the internal standard solution, and the
instrument operation.
41
September 2016

-------
Analytical Protocol for VX Using GC/MS
12.4.4.2 If the above actions do not correct the problem, then the problem
might be due to a sample matrix effect. To determine if there was
matrix effect, take the following corrective action steps:
12.4.4.2.1 Reanalyze the sample extract.
Note: Samples with corresponding MS and MSDs
should be re-extracted and reanalyzed only if
internal standard recoveries in the sample were
considered unacceptable, and the internal standard
recoveries met the acceptance criteria in both the
corresponding MS and MSD.
12.4.4.2.2	If the internal standard responses and RTs meet
acceptance criteria in the re-analyzed sample extract,
then the problem was within the laboratory's control.
12.4.4.2.3	If the internal standard responses and RTs do not meet
acceptance criteria in the re-analyzed sample extract,
flag the results of the associated sample.
12.4.4.2.4	Submit data from both analyses. Distinguish between
the initial analysis and the reanalysis on all data.
13.0 ANALYTICAL PROCEDURE PERFORMANCE
Performance of this protocol was evaluated in multiple laboratories for measurement of VX in
water, soil, and wipe matrices. Resulting IDLs and MDLs from multi-laboratory evaluation are
listed in Table 1, Section 17. Multi-laboratory results of reference matrix samples spiked at
levels corresponding to laboratory low-calibration standards are provided in Table 2, Section 17.
Precision (as RPD and RSD) and bias (as percent recovery) results based on multi-laboratory data
are provided in Table 6, Section 17. Additional laboratory data for real-world samples are
provided in Tables 9a and 9b, Section 17 (multi-laboratory data for groundwater and drinking
water), and 10a - 10b, Section 17 (multi-laboratory data for Virginia-a and ASTM soils).
Characterization data for the samples used during the multi-laboratory study are provided in
Tables 11 and 12, Section 17. Figures 1 and 2 provide example chromatograms generated during
a multi-laboratory study using full-scan quadrupole MS and TOF MS, respectively.
14.0 POLLUTION PREVENTION
14.1 Pollution prevention encompasses any technique that reduces or eliminates the quantity
and/or toxicity of waste at the point of generation. Numerous opportunities for pollution
prevention exist in laboratory operation. EPA has established a preferred hierarchy of
environmental management techniques that places pollution prevention as the
management option of first choice. Whenever feasible, laboratory personnel should use
pollution prevention techniques to address their waste generation. When wastes cannot
42
September 2016

-------
Analytical Protocol for VX Using GC/MS
be feasibly reduced at the source, the Agency recommends recycling as the next best
option.
14.2 For information about pollution prevention that might be applicable to laboratories and
research institutions consult Less is Better: Laboratory Chemical Management for Waste
Reduction, available from the American Chemical Society's Department of Government
Relations and Science Policy, 1155 16th St., N.W. Washington, D.C. 20036, (202) 872-
4477.
15.0 WASTE MANAGEMENT
EPA requires that laboratory waste management practices be conducted in a manner consistent
with all applicable rules and regulations. The Agency urges laboratories to protect the air, water,
and land by minimizing and controlling all releases from hoods and bench operations, complying
with the letter and spirit of any sewer discharge permits and regulations, and by complying with
all solid and hazardous waste regulations, particularly the hazardous waste identification rules
and land disposal restrictions. For further information on waste management, consult The Waste
Management Manual for Laboratory Personnel, available from the American Chemical Society
at the address listed in Section 14.2.
Note: It is strongly recommended that all glassware and waste be decontaminated with bleach
containing active chlorine at a concentration of at least 5 %, for at least 6 hours to provide
effective decontamination of VX. Chemical agent decontamination procedures and inventory
records should be consistent with the laboratory's Chemical Hygiene Plan for chemical warfare
agents.
16.0 REFERENCES
16.1	U.S. Environmental Protection Agency. Selected Analytical Methods for Environmental
Remediation and Recovery (SAM) - 2012. EPA/600/R-12/555. Cincinnati, OH: U.S.
Environmental Protection Agency, Office of Research and Development.
http://www.epa.gov/homeland-securitv-research/sam (accessed 05/31/2016)
16.2	U.S. Environmental Protection Agency. Semivolatile Organic Compounds by Gas
Chromatography/Mass Spectrometry (GC/MS). SW-846 Method 8270D, Revision 4.
February 2007. [In: Test Methods for Evaluating Solid Waste, Physical/Chemical
Methods. EPA publication SW-846. Washington DC: U.S. Environmental Protection
Agency, Office of Land and Emergency Management (formerly, Office of Solid Waste
and Emergency Response).]
16.3	U.S. Environmental Protection Agency. Organic Compounds in Water by Micro
extraction. SW-846 Method 3511, Revision 0. November 2002. Washington DC: U.S.
Environmental Protection Agency, Office of Land and Emergency Management.
16.4	U.S. Environmental Protection Agency. Microscale Solvent Extraction. SW-846
Method 3570, Revision 0. November 2002. Washington DC: U.S. Environmental
Protection Agency, Office of Land and Emergency Management.
43
September 2016

-------
Analytical Protocol for VX Using GC/MS
16.5	U.S. Environmental Protection Agency. Tetra-through Octa-ChlorinatedDioxins and
Furans by Isotope Dilution HRGC/HRMS. Method 1613. October 1994. 62 FR 48394.
Washington, DC: U.S. Environmental Protection Agency, Office of Water.
16.6	Montauban, C., Begos, A., and Bellier, B., Extraction of Nerve Agent VXfrom Soils,
Anal. Chem. 2004, 76, 2791-2797.
16.7	U.S. Army, Marine Corps, Navy and Air Force. Potential Military Chemical/Biological
Agents and Compounds. January 2005. http://fas.org/irp/doddir/army/fm3-ll-9.pdf
(accessed 05/23/2016)
16.8	U.S. Environmental Protection Agency. Semivolatile Organic Compounds in Drinking
Water by Solid-Phase Extraction and Capillary Column (GC/MS), Version 1.0, Method
525.3. February 2012. Cincinnati, OH: U.S. Environmental Protection Agency, Office of
Research and Development. EPA/600/R-12/010.
16.9	U.S. Environmental Protection Agency. Method 1668C: ChlorinatedBiphenyl Congeners
in Water, Soil, Sediment, Biosolids, and Tissue by HRGC/HRMS. Method 1668C. April
2010. Washington DC: U.S. Environmental Protection Agency, Office of Water. EPA-
820-R-10-005.
16.10	U.S. Environmental Protection Agency. "Organic Analytes." Chapter 4 in Test Methods
for Evaluating Solid Waste, Physical/Chemical Methods. EPA publication SW-846.
Washington DC: U.S. Environmental Protection Agency, Office of Land and Emergency
Management.
16.11	ASTM. 2011. Method D1193-06. Standard Specification for Reagent Water, ASTM
International, West Conshohocken, PA. http://www.astm.org (accessed 05/31/2016)
16.12	U.S. Environmental Protection Agency. Stability Study for Ultra-Dilute Chemical
Warfare Agent Standards. EPA 600/R-13/044. May 2013. Cincinnati, OH: U.S.
Environmental Protection Agency, Office of Research and Development, National
Homeland Security Research Center.
16.13	U.S. Environmental Protection Agency. Analytical Standards Requirements. Exhibit E,
Section 4 [In: EPA Contract Laboratory Program Statement of Work for Organic
Superfund Methods Multi-Media, Multi-Concentration SOM02.3, September 2015.
https://www.epa.gov/clp/epa-contract-laboratorv-program-statement-work-organic-
superfund-methods-multi-media-multi-0 (accessed 05/23/2016)]
16.14	U.S. Environmental Protection Agency. Determination of Organic Compounds in
Drinking Water by Liquid-Solid Extraction and Capillary Column Gas
Chromatography/Mass Spectrometry, Method 525.2, Revision 2.0. 1995. Cincinnati, OH:
U.S. Environmental Protection Agency, Office of Research and Development.
16.15	U.S. Environmental Protection Agency. Analytical Standards Requirements. Exhibit C,
[In: Superfund Analytical Services / Contract Laboratory Program Statement of Work for
Multi-Media, Multi-Concentration Organics Analysis (SOM01.2). May 2005.
http: //www. epa.gov/sites/production/file s/2015 -10/documents/som23 d .pdf (accessed
05/31/2016)]
44
September 2016

-------
Analytical Protocol for VX Using GC/MS
16.16 U.S. Environmental Protection Agency. Determinative Chromatographic Separations.
Revision 3, SW-846 Method 8000C. March 2003. Washington DC: U.S. Environmental
Protection Agency, Office of Land and Emergency Management.
45
September 2016

-------
Analytical Protocol for VX Using GC/MS
17.0 TABLES AND FIGURES
Table 1.
Instrument Detection Limits (IDL) and Method Detection Limits (MDL) Based on Multi-
Laboratory Evaluation
Note: IDLs and MDLs were determined in six laboratories during a multi-laboratory study. IDLs were
determined as the concentration necessary to achieve a S:N (signal-noise) ratio of at least 3:1. MDLs
were determined following the procedures in Section 9.7 of this protocol, using spike concentrations
corresponding to the lowest calibration standards.
Sample
Matrix
Full-Scan Quadrupole MS
Full-Scan TOF MS
#
labs
IDL Range
(ng/|jL)(
S:N1
Range
MDL Range
(ra/L)
Pooled
MDL2
(ng/L)
#
labs
MDL Range
(ng/L)
Pooled
MDL2
(ng/L)
Reagent
Water
6
0.04-0.2
-3"
I
cd
0.740-3.06
1.51
6
0.0820-0.44
0.178
Sample
Matrix
#
labs
IDL Range
(ng/pL)
S:N1
Range
MDL Range
(Hg/kg)
Pooled
MDL2
(Hg/kg)
#
labs
MDL Range
(Hg/kg)
Pooled
MDL2
(Hg/kg)
Ottawa
Sand
6
0.04-0.2
-3"
I
cd
1.30-8.283
4.733
6
0.267-0.589
0.328
Sample
Matrix
#
labs
IDL Range
(ng/pL)
S:N1
Range
MDL Range
(|jg/wipe)
Pooled
MDL2
(|jg/wipe)
#
labs
MDL Range
(|jg/wipe)
Pooled
MDL2
(|jg/wipe)
Wipes
6
0.04-0.2
-3"
I
cd
0.0188-
0.0822
0.0457
6
0.00223-0.0129
0.00656
1	S:N value from the weakest of two secondary quantitation ions (see Table 5). S:N values were determined by
measuring peak to peak noise using Agilent Chemstation® software.
2	Pooled MDLs are calculated by taking the square root of the sum of the squares of each laboratory's MDL, divided
by the total number of MDL values and multiplied by a weighting factor based on the degrees of freedom (e.g., for
three MDL values, the weighting factor is 0.81).
3	Includes MDLs resulting from samples spiked at a concentration corresponding to the second lowest calibration
standard.
46
September 2016

-------
Analytical Protocol for VX Using GC/MS
Tables 2a and 2b: Multi-Laboratory Results of Reference Matrix Samples Spiked at Levels
Corresponding to Laboratory Low-Calibration Standards
Note: Table 2 provides summary results of reference matrix samples spiked at levels corresponding to
the lowest calibration standard used by laboratories participating in the multi-laboratory protocol validation
study. The last column of this table provides spike levels adjusted based on the lowest recovery result,
for comparison to pooled MDLs generated using study data. With the exception of the analysis of VX in
wipes using GC-TOF, the adjusted spike levels are above the pooled MDLs in all cases.
Table 2a.
Example Multi-Laboratory Results for Reference Matrix Samples Analyzed Using
Full-Scan Quadrupole MS
# of
labs
n
Spike Level
%Recovery
Range
Pooled
RSD
Pooled MDL
Spike Level
adjusted for
recovery
Reagent Water (|jg/L)()) ()
6
42
5.71
53.3 - 264
10.3 1 1.51
3.05
Ottawa Sand (pg/kg) )
4
28
10.0
62.0-283
13.2 | 4.73*
6.20
Wipes (|jg/wipe)
5
35
0.100
79.0-245
11.8 | 0.457
0.0790
n = number of replicates; MDL = method detection limit; RSD = relative standard deviation
includes results of samples spiked at levels corresponding to the second lowest calibration standard
Table 2b.
Example Multi-Laboratory Results for Reference Matrix Samples Analyzed Using
Full-Scan TOF MS
# of labs
n
Spike Level
%Recovery
Range
Pooled Pooled
RSD MDL
Spike Level adjusted
for recovery
Reagent Water (|jg/L)
6
42
0.571
60.6-164
10.9 1 0.178
0.346
Ottawa Sand (|jg/kg)
4
28
1.00
61.0-250
9.80 | 0.328
0.610
Wipes (|jg/wipe)
4
28
0.0100
0 -188
87.4 | 0.00656
0*
n = number of replicates; MDL = method detection limit; RSD = relative standard deviation
*Ten of 28 recoveries produced an adjusted spike level that is lower than the pooled MDL
47
September 2016

-------
Analytical Protocol for VX Using GC/MS
Table 3.
Decafluorotriphenylphosphine (DFTPP) Key Ions and Ion Abundance Criteria
Note: All ion abundances MUST be normalized to m/z 198.
Mass
Quadrupole
Time of Flight (TOF)
51
10.0 - 80.0 % of mass 198
10.0 - 85.0 % of mass 198
68
Less than 2.0 % of mass 69
Less than 2.0 % of mass 69
69
Present
Not used
70
Less than 2.0 % of mass 69
Less than 2.0 % of mass 69
127
10.0 - 80.0 % of mass 198
10.0 - 80.0 % of mass 198
197
Less than 2.0 % of mass 198
Less than 2.0 % of mass 198
198
Base peak 100 % relative abundance
(see Note above)
Base peak 100 % relative abundance
199
5.0 - 9.0 % of mass 198
5.0 - 9.0 % of mass 198
275
10.0 - 60.0 % of mass 198
10.0 - 60.0 % of mass 198
365
Greater than 1.0 % of mass 198
Greater than 0.5 % of mass 198
441
Present but less than mass 443
Less than 150 % of mass 443
442
Greater than 50.0 % of mass 198
Greater than 30.0 %
443
15.0 - 24.0 % of mass 442
15.0 - 24.0 % of mass 442
Table 4.
Internal Standard and Surrogates
Note: Table 4 provides a list of surrogates used during laboratory studies testing this protocol,
based on surrogates typically used for SVOC analyses. Alternative surrogates might be more
representative of VX, and may be used or added at the laboratory's discretion provided the
surrogates meet the criteria in Table 7.
Analyte
Surrogate Compound
Internal Standard
VX
Triphenyl phosphate
Phenanthrene-dio
Table 5.
Example Retention Times, Relative Retention Times and Quantitation Ions for Target
Compounds, Surrogate Compounds, and Internal Standard
Note: Bold quantitation ions indicate the secondary ions used during single-laboratory testing.
Analyte
Retention
Time (min)
Relative
Retention
Time
Primary
Quantitation Ion
Secondary
Quantitation Ions
VX
17.50-17.51
0.97
114
127, 79, 72
Triphenyl phosphate (S)
20.68
1.64
326
325, 215
Phenanthrene-dio (IS)
18.02-18.03
-
188
94, 80
(S) = Surrogate
(IS) = Internal Standard
48
September 2016

-------
Analytical Protocol for VX Using GC/MS
Table 6.
Example Multi-Laboratory Precision and Bias in Reference Matrices
at Mid-Calibration Levels
Sample
Matrix
Initial Precision and
Recovery
(IPR)
Laboratory
Control Sample
(LCS)
Matrix Spike/Matrix Spike
Duplicate (MS/MSD)
%
Recovery
Pooled
RSD
%
Recovery
%
Recovery
Pooled
RPD
Full-Scan Quadrupole MS
Water1
61.0-121
10.3
61.0-121
61.0-121
14.6
Soil2
51.8-142
12.1
51.8-142
51.8-142
17.2
Wipes3
45.9-154
13.8
45.9-154
45.9-154
19.5
Full-Scan TOF MS
Water1
49.0-154
10.2
49.0-154
49.0-154
14.4
Soil2
30.0-131
22.9
30.0-131
30.0-131
32.44
Wipes3
29.3-125
16.5
29.3-125
29.3-125
23.4
1	Reagent water spike levels: 28.6 - 57.1 |jg/L (Full Scan); 5.71 - 22.86 |jg/L (TOF)
2	Soil spike levels: 40.0 - 80.0 [jg/kg (Full Scan); 10-40 [jg/kg (TOF)
3	Wipe spike levels: 0.4 - 1.0 pg/wipe (Full Scan); 0.1 - 0.4 pg/wipe (TOF)
4	Removal of one of 20 results from this dataset yields a pooled RPD of 21.0 %
Table 7.
Surrogate Recovery
Note: Table 7 list the surrogate used during laboratory studies testing this protocol. Alternative
surrogates might be more representative of VX, and may be used or added at the laboratory's
discretion, provided the surrogates meet the criteria listed in this table.
Sample Matrix
Surrogate
Surrogate %Recovery
Minimum
Maximum
Full-Scan Quadrupole MS
Water
Triphenyl phosphate
50.0
150
Soil
Triphenyl phosphate
50.0
150
Wipes
Triphenyl phosphate
50.0
150
Full-Scan TOF MS
Water
Triphenyl phosphate
50.0
150
Soil
Triphenyl phosphate
50.0
150
Wipes
Triphenyl phosphate
50.0
150
49
September 2016

-------
Analytical Protocol for VX Using GC/MS
Table 8.
Example Calibration Standard Concentrations (|jg/mL)
Used During Multi-Laboratory Method Validation Study
Full-Scan Quadrupole MS

Analyte
CAS RN
Cal 1
Cal 2
Cal 3
Cal 4
Cal 5
Cal 6
VX
50782-69-9
0.1
0.2
0.4
0.8
1.0
2.0
Triphenyl phosphate (S)
115-86-6
0.1
0.2
0.4
0.8
1.0
2.0
Phenanthrene-dio(IS)
1517-22-2
0.5
0.5
0.5
0.5
0.5
0.5
Full-Scan TOF MS
Analyte
CAS RN
Cal 1
Cal 2
Cal 3
Cal 4
Cal 5
Cal 6
Cal 7
VX
50782-69-9
0.01
0.05
0.08
0.1
0.25
0.5
1.0
Triphenyl phosphate (S)
115-86-6
0.01
0.05
0.08
0.1
0.25
0.5
1.0
Phenanthrene-dio(IS)
1517-22-2
0.5
0.5
0.5
0.5
0.5
0.5
0.5
(S) = Surrogate; (IS) = Internal standard
50
September 2016

-------
Analytical Protocol for VX Using GC/MS
Table 9a.
Example Multi-laboratory Precision and Recovery in Water
Using GC Full-Scan Quadrupole MS
Groundwater
Analyte
# of labs
n
Pooled
%RSD
%Recovery
Range
%RSD
Range
VX (5.71 |jg/L)
3
21
8.02
58.8-133
4.06-12.4
VX (45.7 |jg/L)
3
21
5.58
62.4-102
3.70-7.65
Surrogate
Triphenylphosphate (28.6 pg/L)
3
42
9.45
60.0-156
4.17-18.1
Drinking Water
Analyte
# of labs
n
Pooled
%RSD
%Recovery
Range
%RSD
Range
VX (5.71 |jg/L)
3
21
26.9
58.1 - 125
5.77-44.7
VX (45.7 |jg/L)
3
21
5.76
60.8-108
3.65-7.41
Surrogate
Triphenylphosphate (28.6 pg/L)
4
70
7.44
52.0-124
2.01 - 12.6
n = number of replicates; RSD = relative standard deviation
Table 9b.
Example Multi-laboratory Precision and Recovery in Water
Using GC Full-Scan TOF MS
Groundwater
Analyte
# of
labs
n
Pooled
%RSD
%Recovery
Range
%RSD
Range
VX (0.571 |jg/L)
3
21
28.4
11.8-106
4.86-48.0
VX (5.71 |jg/L)
3
21
3.38
41.7-122
4.96-23.6
Surrogate
Triphenylphosphate (28.6 pg/L)
4
64
7.82
36.6-143
2.35-18.6
Drinking Watter
Analyte
# of
labs
n
Pooled
%RSD
% Recovery
Range
%RSD
Range
VX (0.571 |jg/L)
3
21
15.9
28.9-107
5.63-23.5
VX (5.71 |jg/L)
3
21
6.74
72.6-117
2.53 -10.6
Surrogate
Triphenylphosphate (28.6 pg/L)
4
62
7.23
69.0-136
2.81 - 16.9
n = number of replicates; RSD, relative standard deviation; TOF, time of flight
51
September 2016

-------
Analytical Protocol for VX Using GC/MS
Table 10a.
Example Multi-laboratory Precision and Recovery in Soils
Using GC Full-Scan Quadrupole MS
Virginia-a Soil
Analyte
# of
labs
n
Pooled
%RSD
%Recovery
Range
%RSD
Range
VX(10.0 [jg/kg)
3
21
8.60
80.5-144
1.47-13.8
VX (80 pg/kg)
3
21
13.3
19.2-68.2
9.56-16.8
Surrogate
Triphenylphosphate (50 pg/kg)
3
31
29.0
0.516-76.0
6.43-55.8

ASTM Soil
Analyte
# of
labs
n
Pooled
%RSD
% Recovery
Range
%RSD
Range
VX(10.0 pg/kg)
3
21
2.35
66.8-113
1.71 -2.80
VX (80.0 pg/kg)
3
20
16.1
5.60-42.9
6.17-21.5
Surrogate
Triphenylphosphate (50 pg/kg)
3
31
16.1
5.57-194
5.87-27.6
n = number of replicates; RSD, relative standard deviation
Table 10b.
Example Multi-laboratory Precision and Recovery in Soils
Using GC Full-Scan TOF MS
Virginia -a Soil
Analyte
# of
labs
n
Pooled
%RSD
%Recovery
Range
%RSD
Range
VX (1.0 pg/kg)
2
14
11.7
56.9-146
6.14-15.3
VX(10.0 pg/kg)
3
18
8.12
34.0-84.6
7.37-9.11
Surrogate
Triphenylphosphate (50 pg/kg)
3
39
14.7
2.32-7.82
7.77-16.1
ASTM Soil
Analyte
# of
labs
n
Pooled
%RSD
%Recovery
Range
%RSD
Range
VX (1.0 pg/kg)
0
0
NA
0
NA
VX(10.0 pg/kg)
3
16
20.7
12.7-34.0
7.74-34.0
Surrogate
Triphenylphosphate (50 pg/kg)
3
39
18.8
9.31 -47.6
3.68-30.6
n = number of replicates; RSD = relative standard deviation; TOF = time of flight
52
September

-------
Analytical Protocol for VX Using GC/MS
Table 11.
Multi-Laboratory Study Water Matrices Characterization Data
Drinking Water
Groundwater
Collection Date
Dec 2011 -
July 2012
Collection Date
Jan/Feb 2012
PH
00
cn
I
00
PH
6-7
Chlorine (free) mg/L
1.16-1.35
Chlorine (mg/L)
0.02-0.05
Chlorine (total) mg/L
1.20-1.42
Chlorine (total) mg/L
NR
Total Organic Carbon (TOC) (ppm)
0.83-1.14
Total Organic Carbon (TOC) (ppm)
1.08-1.24
Conductivity (|jS)
294-413
Conductivity (|jS)
0.82-0.87
Oxidation-reduction potential (mV)
766 - 770
Oxidation-reduction potential (mV)
NR
Turbidity (NTU)
0.06-0.07
Turbidity (NTU)
CO
o
I
o
Total Hardness (mg/L)
114-131
Total Hardness (mg/L)
37-65
Alkalinity (mg/L)
64-80
Alkalinity (mg/L)
30-36
Table 12.
Multi-laboratory Study Soil Matrix Characterization Data
Parameter

ASTM Soil ML-1
Virginia-a Soil
PH
8.68
4.41
TOC (%C)
<0.10
2.2
%Solid
98.9
99.1
TOC = total organic carbon
53
September 2016

-------
Analytical Protocol for VX Using GC/MS
Abundance
600000
500000
400000
300000
200000
100000
x
>
¦o
i
Time (min) 16.00
V
18.00
r.
a.

20.00
22.00
Notes:
(1)	Concentrations of all analytes as described in Table 8, Calibration Level 6 (GC/MS - Full Scan Quadrupole);
(2)	Unlabelled peaks represent compounds not specifically targeted by this protocol
(min) = minutes
I = internal standard
S = surrogate
T = target
Figure 1.
Example chromatogram for a calibration standard on full-scan quadrupole MS.
54
September 2016

-------
Analytical Protocol for VX Using GC/MS
x
>
CD
•*->
CO
£
Q.
(/)
O
>*
c
CD

Aj
175
200
I
225
250
Time(s) 150
Notes:
(1)	Concentrations of all analytes as described in Table 8, Calibration Level 4 (GC/MS -TOF);
(2)	Unlabelled peaks represent compounds not specifically targeted by this protocol
(s) = seconds
275
300
—t—-
325

375
Figure 2.
Example chromatogram for a midpoint calibration standard (Cal5) on time-of-flight MS.
55
September 2016

-------
vvEPA
United States
Environmental Protection
Agency
PRESORTED STANDARD
POSTAGE & FEES PAID
EPA
PERMIT NO. G-35
Office of Research and Development (8101R)
Washington, DC 20460
Official Business
Penalty for Private Use
$300

-------