OBaaefle
- - - Putting Technology To Work
Environmental Technology
Verification Program
Advanced Monitoring
Systems Pilot
Test/QA Plan for Verification
of Optical Open-Path
Monitors
ET^TWIV

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 1 of 67
TEST/QA PLAN
FOR
VERIFICATION OF
OPTICAL OPEN-PATH MONITORS
October 28, 1999
Prepared by
Battelle
505 King Avenue
Columbus, OH 43201-2693

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 2 of 67
TEST/QA PLAN FOR VERIFICATION OF
OPTICAL OPEN-PATH MONITORS
October 28,1999
APPROVAL
Vendor Name	Approval
Date

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 3 of 67
Table of Contents
Page
1.	Introduction	6
1.1.	Test Description	6
1.2.	Test Objective	6
1.3.	Applicability	7
2.	Technology Description	7
3.	Verification Approach	8
3.1.	Scope of Test 	8
3.2.	Organization and Responsibilities 	8
3.2.1	Battelle 	10
3.2.2	Vendors	12
3.2.3	EPA 	13
4.	Experimental Design	14
4.1.	Overview	14
4.2.	General Experimental Method 	14
5.	Test Procedures	15
5.1.	General Procedural Description	15
5.2.	Schedule 	17
5.3.	I TIR	19
5.3.1	Gases 	19
5.3.2	Minimum Detection Limit	19
5.3.3	Linearity	21
5.3.4	Accuracy	22
5.3.5	Precision	23
5.3.6	Interferences 	23
5.4.	Tuneable Diode Laser 	24
5.4.1	Gases 	24
5.4.2	Minimum Detection Limit	25
5.4.3	Linearity	25
5.4.4	Accuracy	26
5.4.5	Precision	27
5.4.6	Interferences 	27
5.5.	Ultra Violet (UV) Open Path Monitors	28
5.5.1 Gases 	28

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 4 of 67
5.5.2	Minimum Detection Limit	28
5.5.3	Linearity	30
5.5.4	Accuracy	31
5.5.5	Precision	31

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 5 of 67
Table of Contents (Continued)
Page
5.5.6 Interferences 	32
6.	Site Description	33
7.	Materials and Equipment 	33
7.1.	Standard Gases	33
7.2.	Dilution Gas	33
7.3.	Dilution System	34
7.4.	Temperature Sensor	34
7.5.	Relative Humidity Sensor	34
7.6.	Oxygen Sensor 	34
7.7.	Carbon Dioxide Sensor	35
7.8.	Ozone Sensor 	35
7.9.	Monitor for NO and NH3	35
7.10.	Carbon Monoxide Sensor 	35
8.	Quality Assurance/Quality Control	37
8.1.	Calibration	37
8.1.1	Gas Dilution System	37
8.1.2	Temperature Sensor	37
8.1.3	Relative Humidity Sensor	37
8.1.4	Oxygen Sensor 	38
8.1.5	Carbon Monoxide Sensor 	38
8.1.6	Carbon Dioxide Sensor	38
8.1.7	Ozone Sensor 	39
8.1.8	Monitor for NO and NH3	39
8.2.	Assessments and Audits	40
8.2.1	Technical Systems Audit	40
8.2.2	Performance Evaluation Audits	40
8.2.3	Data Quality Audit	42
8.3.	Assessment Reports	43
8.4.	Corrective Action	43
9.	Data Analysis and Reporting	43
9.1.	Data Acquisition	43
9.2.	Statistical Calculations 	44
9.2.1 Minimum Detection Limit	45

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 6 of 67
9.2.2	Linearity	45
9.2.3	Accuracy	46
9.2.4	Precision	46
9.2.5	Interferences 	47

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 7 of 67
Table of Contents (Continued)
Page
9.3.	Data Review 	47
9.4.	Reporting	48
10.	Health and Safety	48
10.1.	General	48
10.2.	Potential Hazards	49
10.3.	Training	49
11.	Definitions 	50
12.	References	52
APPENDDC A - Example Data Sheet 	53
List of Figures
1.	Organizational chart for optical open-path monitor verification test	9
2.	Schematic showing functional system and setup for verification tests	16
List of Tables
1.	Optical Open-Path Monitor Verification Measurement Order for a Single Gas	18
2.	Schedule of Verification Testing Activities 	19
3.	Target Gases for Verification Testing of FTIR Open-Path Monitors	20
4.	Target Gases for Verification Testing of TDL Open-Path Monitors	24
5.	Target Gases for Verification Testing of UV Open-Path Monitors 	29

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 8 of 67
6.	Summary of Performance Evaluation Audit Procedures 	41
7.	Summary of Data Recording Process	45
1. Introduction
1.1. Test Description
This test/QA plan provides detailed procedures for a verification test of optical open-path
monitors for use in ambient air or fenceline measurements. The verification test will be conducted under
the auspices of the U.S. Environmental Protection Agency (EPA) through the Environmental
Technology Verification (ETV) program. The purpose of ETV is to provide objective and quality
assured performance data on environmental technologies, so that users, developers, regulators, and
consultants have an independent and credible assessment of what they are buying and permitting.
The verification test will be performed by Battelle, of Columbus, Ohio, which is managing the
ETV Advanced Monitoring Systems (AMS) pilot through a cooperative agreement with EPA
(CR826215). The scope of the AMS pilot covers verification of monitoring methods for contaminants
and natural species in air, water, and soil. In performing the verification test, Battelle will follow
procedures specified in this test/QA plan, and will comply with the quality requirements in the "Quality
Management Plan for the ETV Advanced Monitoring Systems Pilot" (QMP).1
1.2. Test Objective
The goal of this verification test is to quantify the verification parameters of commercially
available optical open-path monitors for use at facilities concerned with emissions or ambient levels of
volatile organic or inorganic chemicals. This verification will involve challenging these monitors with
known reference gas samples under realistic operational conditions.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 9 of 67
1.3. Applicability
The technologies tested under this plan are commercial optical open-path monitors capable of
real-time monitoring of atmospheric pollutants. These monitors are typically used over greater than 100
meter path lengths, and present the users with information about the concentrations of gases that are
present in the air between the light source and the detector. In such applications these open-path
monitors can provide real-time continuous monitoring of air quality, and allow early warning of potential
non-compliance conditions, or emergency release situations. In contrast, grab sample analysis by
standard methods is both time-consuming and non-continuous.
2. Technology Description
The monitors to be verified under this test/QA plan rely upon a radiation source (ultraviolet,
visible, or infrared) and a detector used together to identify and quantify the levels of certain chemicals
in the atmosphere. These monitors are typically used in a continuous monitoring mode and in many
cases are able to provide simultaneous monitoring of several compounds. Although the overall design
requirements for the different spectral ranges are significantly different, the basic components of these
technologies are similar.
In general, these monitors contain at least the following components:
Radiation source
Optics
Detector
Data processing algorithms.
The radiation sources for these technologies belong to one of three distinct groups. The
monitors operating in the ultraviolet (UV) region of the spectrum use a continuous or non-continuous

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 10 of 67
lamp that provides broad-band radiation in the UV and visible regions. The monitors using tunable
diode laser (TDL) technology use a laser to provide radiation over a very narrow spectral range in the
near infrared. That spectral range can be tuned over a small range with a single TDL, and is selectable
over a wider range using multiple TDLs. The Fourier transform infrared (FTIR) monitors use a
broadband infrared source.
The optical components of these monitors typically are used to project the radiation from the
source, through the atmospheric path to be monitored, and to the detector. The detectors and
configurations for these monitors vary according to specific applications. They are typically chosen to
maximize signal to noise ratio for the spectral region and operating temperature.
3. Verification Approach
3.1. Scope of Test
The overall objective of the verification test is to provide quantitative verification of the
performance of optical open-path monitors under realistic operational conditions. Specifically, the
verification parameters to be verified are:
•	Minimum detection limit (MDL)
Concentration linearity
Source strength linearity
Accuracy
•	Precision
Sensitivity to atmospheric interferences.
3.2. Organization and Responsibilities

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 11 of 67
The verification test will be performed by Battelle with the participation of EPA and the vendors
whose optical open-path monitors will be verified. The organizational chart in Figure 1 shows the
individuals from Battelle, the vendor companies, and EPA who will have responsibilities in the
verification test. The specific responsibilities of these individuals are detailed below.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 12 of 67
Sandy Anderson
Quality Manager
Battelle
Management
Karen Riggs
Pilot Manager
Thomas Kelly
Verification
Test Leader
Jeffrey Myers
Verification
Test Coordinator
Robert Fuerst
EPA Pilot
Manager
Elizabeth Betz
EPA Pilot
Quality Manager
\
Open-Path

Vendor

Representatives
Battelle
Testing
Staff

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 13 of 67
Figure 1. Organizational Chart for Optical Open-Path Monitor Verification Test

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 14 of 67
3.2.1 Battelle
Mr. Jeffrey D. Myers is the Verification Test Coordinator for the testing of optical open-path
monitors through the AMS pilot. In this role, Mr. Myers will have the overall responsibility for ensuring
that the technical, scheduling, and cost goals established for the verification test are met. Mr. Myers
will:
Prepare the draft test/QA plan, verification reports, and verification statements
Revise the draft test/QA plan, verification reports, and verification statements inresponse to
reviewers' comments
Assemble the requisite equipment and a team of qualified technical staff to conduct the
verification test
Direct the team in performing the verification test in accordance with the test/QA plan
Coordinate distribution of the final test/QA plan, verification reports and verification
statements
Ensure that all quality procedures specified in the test/QA plan and in the QMP are
followed
Respond to any issues raised in assessment reports and audits, including instituting
corrective action as necessary
Serve as the primary point of contact for vendor representatives
Establish a budget for the verification test and monitor staff effort to ensure that the budget
is not exceeded
Ensure that confidentiality of vendor information is maintained.
Dr. Thomas J. Kellv is the Verification Testing Leader for the AMS pilot. As such, Dr. Kelly
will provide technical guidance and oversee the various stages of the verification test. He will:

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 15 of 67
Support Mr. Myers in preparing the test/QA plan and organizing the test
Review the draft test/QA plan
Review the draft verification reports and verification statements
Ensure that vendor confidentiality is maintained.
Ms. Karen Riggs is Battelle's ETV pilot manager. As such, Ms. Riggs will:
Review the draft test/QA plan
Review the draft verification reports and verification statements
Ensure that necessary Battelle resources, including staff and facilities, are committed to the
verification test
Ensure that vendor confidentiality is maintained
Support Mr. Myers in responding to any issues raised in audits
• Maintain communication with EPA's Pilot Manager.
Ms. Sandv Anderson will serve as the Quality Manager for this verification test. As such Ms.
Anderson or her designate will:
Review the draft test/QA plan
Conduct a technical system audit once during the verification test
Review performance evaluation audit results as specified in the test/QA plan
Audit at least 10% of the verification data
Prepare and distribute an assessment report for each audit
Verily implementation of any necessary corrective action

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 16 of 67
Issue a stop work order if self audits indicate that data quality is being compromised; notify
Battelle Pilot Manager if stop work order is issued
Provide a summary of the QA/QC activities and results for the verification reports
Review the draft verification reports and statements
Have overall responsibility for ensuring that the test/QA plan and AMS pilot QMP is
followed
Ensure that Battelle management is informed if persistent quality problems are not
corrected
• Maintain communication with EPA's Pilot Quality Manager
3.2.2 Vendors
Vendor representatives will:
Review the draft test/QA plan
Approve the revised test/QA plan
Provide off-the-shelf models of the optical open-path monitors to be verified for the
duration of the verification test
Host verification testing of their monitors at their respective facilities, or send monitor and
personnel to Battelle to conduct test
Install the test equipment and open-path monitor in the test facilities and ensure proper
operation of the open-path monitor before and during the test

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 17 of 67
Perform on-site maintenance as necessary if monitor fails any time during the test
Review their respective draft verification reports and statements
Provide measurement results from the verification test to Battelle in a readily accessible and
previously agreed upon format
Provide and operate the open-path monitor during testing
Provide sample gas cell appropriate for the monitor being tested
If test is performed at Battelle, remove and ship monitor from Battelle upon completion of
test
3.2.3 EPA
EPA's responsibilities in the AMS pilot are based on the requirements stated in the
"Environmental Technology Verification Program Quality and Management Plan of the Pilot Period
(1995-2000)" (QMP). The roles of the specific EPA staff are as follows:
Ms. Elizabeth Betz is EPA's Pilot Quality Manager. For the verification test, Ms. Betz will:
Review the draft test/QA plan
Perform, at her option, one external technical system audit during the verification test
• Notify the Battelle Pilot Manager to facilitate a stop work order if external audit indicates
that data quality is being compromised
Prepare and distribute an assessment report summarizing results of external audit, if
performed

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 18 of 67
Review draft verification reports and statements.
Mr. Robert Fuerst is EPA's Pilot Manager. As such, Mr. Fuerst will:
Review the draft test/QA plan
Approve the final test/QA plan
Review the draft verification reports
Approve the final verification reports
Review the draft verification statements.
4. Experimental Design
4.1. Overview
The verification test outlined in this document is designed to challenge the monitors being
verified in a manner similar to that which would be experienced in operation in the field. Reproducing
many of the actual conditions and problems encountered in the field is beyond the scope of this project,
however this verification test establishes benchmarks that provide quantitative data on specific
performance parameters. The basic theory used throughout these tests involves challenging the
monitors using an optically transparent gas cell that is filled with known concentrations of a target gas.
The gas cell is inserted into the optical path of the monitor thereby simulating a condition where the
target gas would be present in the ambient air. The gas cell is used to challenge the monitor in a
controlled and uniform manner.
4.2. General Experimental Approach

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 19 of 67
This verification test is intended to be applicable to many types of open-path monitors. As
such, the general approach is deliberately broad, allowing specific protocols for a technology type to be
specified in the Test Procedures section. In general, the experimental approach employed in this test
assumes that the monitor operates by sending a beam of radiation from a source, through the
atmosphere, and to a detector. Then, measuring the absorption of the light by the target gas in the
atmosphere, the monitor is able to identify and quantify the target gas or gases. The same basic
technique is used to verify each of the technology types as each of the monitors is challenged with
several target gases at known concentrations, and the measurement result from the monitor is then
compared to the known concentration of the target gas. Since these open-path monitors are often able
to measure many different types of gases it is not feasible to test all potential target gases. As a result,
in this verification test only a few target gases have been chosen. For each target gas the monitor is set
up as it would be if it were operating in the field, with the exception that an optically transparent gas cell
is placed in the light beam's path if the instrument does not already have a built-in gas reference cell. A
known concentration of the target gas is then introduced into the gas cell and the monitor makes a
measurement. Figure 2 shows a schematic of the setup to be used for the test. The optical open-path
monitor and the gas cell will be provided by the vendor. The gas dilution system will be provided by
Battelle. This system consists of National Institute of Standards and Technology (NIST) traceable
commercially certified standard gases, a calibrated gas diluter, and a supply of certified high purity
dilution gas. All of the test equipment used to evaluate the monitors will be provided by Battelle to
ensure that the testing is conducted in a repeatable manner regardless of the test location. When testing
is performed at a site other than Battelle, all appropriate equipment will be transported to the test site.
The test procedures involve providing a range of known concentrations of various target gases
to each monitor. Measurements are made with different path lengths, integration times, source
intensities, and numbers of replicate measurements to assess the verification parameters listed in Section
3.1. The test procedures are 'nested', in that each measurement is used for the evaluation of more than
one verification parameter. To the extent feasible with so diverse a group of technologies, verification

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 20 of 67
test procedures rely on established procedures, such as EPA Method TO-16.2 This method was
developed to provide guidelines for gathering and analyzing data using an FTIR.
5. Test Procedures
5.1. General Procedural Description
The procedures to be used in the verification test are detailed in this section. This test
procedure section is divided into three sub sections - FTIR, TDL, and UV - each outlining the specific
test procedures required for a particular technology type.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 21 of 67
Gas Reference Cell
Optical path
Dilution
gas
Target gas
or gases
To
Vent
Optical
Open-Path
Monitor
Gas Dilution
System
Figure 2. Schematic Showing Functional System and Setup for Verification Tests

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 22 of 67
The procedures detailed in this section can be carried out for many different target gases. In an
effort to be as efficient as possible with both time and materials a specific order of measurements has
been established which allows many of the verification parameters to be determined in as short a period
as possible. Table 1 shows the measurement order and the
verification parameters associated with each measurement. The "Activity #" column provides a
reference number for each activity during the test. This allows for easy reference later in the test plan.
The "Meas. #" column is listed to show the number of times spectra are recorded. "Ref. Cell Cone."
describes the contents of the reference gas cell during the data acquisition. In these measurements, the
content of the gas cell is either "N2" (nitrogen dilution gas), or a known concentration of the target gas
"cl" through "c4". Concentrations shown as cl, c2, etc. represent the target gas concentrations
specified later in this section for each technology type. The "Activity" column explains the activity
taking place: collecting spectra, changing gases, or adjusting the pathlength. Several measurements are
made (Meas. #3 through #5 and #10 through #12) using an inserted neutral density (ND) filter which
allows the source strength (i.e. light intensity) to be varied in a controlled and repeatable manner. The
"# of Spectra" column explains how many individual spectra are collected at that experimental
condition. The "Integrate Time" column is the integration time to be used for that measurement and
"Equilibrate Time" expresses the length of time allowed to flow gas through the cell in order to obtain a
stable measurement from the monitor. "Pathlength" is the total length the beam will travel between the
source and the detector. This is not the length of the gas cell used in these experiments. From the
"Elapsed Time" column it can be seen that the test will require about 8.25 hours for the first gas. The
subsequent gases take an half an hour less since the replicate tests for the ND filter are not conducted.
The "Verification Parameter Calculated" column relates each measurement to the verification
parameters that will eventually be calculated.
5.2. Schedule

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 23 of 67
The verification will be conducted by performing measurements in a fixed sequence. The
monitor provided by the vendor will undergo that full test sequence. It is anticipated that the

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 24 of 67
Table 1. Optical Open-Path Monitor Verification Measurement Order for a Single Gas
Activity
Vleas
Kef. Cell
Activity
#of
Times
Pathlength
Elapsed
Verification Parameter
#
#
Cone.

Spectra
Integrat
e
Equilibrat
(m)
Time (hrs)
Calculated
1

N2
Change gas & stabilize


10
100
1/4

2
1
N2
Collect spectra
26
1

100
2/4
Accuracy., Concentration linearity, MDL, Precision
3

c 1
Change gas & stabilize


10
100
3/4

4
2
c 1
Collect spectra
5
1

100
3/4
Acc., Concentration linearity
5
3
c 1
Collect spectra - ND 1
5
1

100
1
Source strength linearity
6
4
c 1
Collect spectra - ND 2
5
1

100
1
Source strenath linearity
7
5
c 1
Collect spectra - ND 3
5
1

100
1
Source strength linearity
8

N2
Change gas & stabilize


10
100
1 1/4

9
6
N2
Collect spectra
5
1

100
1 1/4
Acc., Concentration linearity
10

c2
Change gas & stabilize


10
100
1 2/4

11
/
c2
Collect spectra
5
1

100
1 2/4
Acc., Concentration linearity
12

N2
Change gas & stabilize


10
100
1 3/4

13
8
N2
Collect spectra
5
1

100
1 3/4
Acc., Concentration linearity
14

c3
Change gas & stabilize


10
100
2

1b
9
c3
Collect spectra
5
1

100
2
Acc.. Concentration linearity
16
10
c3
Collect spectra - ND 1
5
1

100
2 1/4
Source strength linearity
17
11
c3
Collect spectra - ND 2
5
1

100
2 1/4
Source strength linearity
18
12
c3
Collect spectra - ND 3
5
1

100
2 1/4
Source strenath linearity
19

N2
Change gas & stabilize


10
100
2 2/4

20
13
N2
Collect spectra
5
1

100
2 2/4
Acc., Concentration linearity
21

c4
Change gas & stabilize


10
100
2 3/4

22
14
c4
Collect spectra
26
1

100
2 3/4
Acc., Concentration linearity, MDL, Precision
23

N2
Change gas & stabilize


10
100
3

24
15
N2
Collect spectra
26
5

100
5 1/4
Acc.. Concentration linearity. MDL. Precision
25


Change to Pathlength 2


20
400
5 2/4

26
16
N2
Collect spectra
5
5

400
6
Interference Effect (Int.)
27

c2
Change gas & stabilize


10
400
6

28
1/
c2
Collect spectra
5
5

400
6 2/4
Int., Acc., Concentration linearity
29

N2
Change gas & stabilize


10
400
6 3/4

30
18
N2
Collect spectra
5
5

400
/
Int., Acc., Concentration linearity
31


Change to Pathlength 3


20
optimum
7 2/4

32
19
N2
Collect spectra
5
1

optimum
7 214
Int., Acc., Concentration linearity
33

c2
Change gas & stabilize


10
optimum
7 3/4

34
20
c2
Collect spectra
5
1

optimum
/ 3/4
Int., Acc., Concentration linearity

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 25 of 67
35

N2
Change gas & stabilize


10
optimum
§

36
21
N2
Collect SDectra
26
1

ODtimum
8 1/4
Int.. Precision. MDL

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 26 of 67
testing will take place over three days, with a day for set up and a day for teardown. An example
schedule of testing a single monitor is shown in Table 2.
Table 2. Schedule of Verification Testing Activities
Activity #
Day
Gas
Approximate Time
Travel and Setup
One
—
1 day
1-24
Two
One
08:00-13:00
25-36
Two
One
14:00-18:00
1-24
Three
Two
08:00-12:00
25-36
Three
Two
13:00-17:00
1-24
Four
Three
08:00-12:00
25-36
Four
Three
13:00-17:00
Teardown and Travel
Five
—
1 day
5.3. FTIR
5.3.1 Gases
The gases and concentrations to be used for testing FTIR open-path monitors are shown in
Table 3.
5.3.2. Minimum Detection Limit
The minimum detection limit (MDL) is to be determined for each target gas. This number
represents the lowest obtainable value for the detection of that specific gas. The MDL is calculated by

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 27 of 67
removing the target gas from the optical path of the monitor, then a series of 26 single beam spectra are
taken using the appropriate averaging time (either 1 min. or 5 min.). The
Table 3. Target Gases for Verification Testing of FTIR Open-Path Monitors
Gas

Concentration
Pathlength (ppm-m)

cl
5
T etrachloroethylene
c2
10

c3
25

c4
50

cl
5
Cyclohexane
c2
10

c3
25

c4
50

cl
5
Ethylene
c2
10

c3
25

c4
50
single beam spectra are then used to create absorption spectra, using each single beam spectrum as the
background for the next spectrum. The absorption spectra are created by using the first and second
single beam spectra, the second and third, the third and fourth, etc. The resulting 25 absorption spectra
are then analyzed for the target gas. The minimum detection limit is defined as two times the standard
deviation of the calculated concentrations.
The procedure for determining MDL is as follows:

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 28 of 67
1.	Remove the target gas from the optical path of the monitor
2.	Choose appropriate averaging time for the monitor
3.	Acquire 26 single beam spectra
4.	Use first single beam spectrum as a background to create absorption spectrum from the
second single beam spectrum
5.	Use second single beam spectrum as a background to create absorption spectrum from the
third single beam spectrum
6.	Continue until 25 absorbance spectra are obtained
7.	Analyze each absorption spectrum to determine the concentration of the target gas
8.	Calculate the standard deviation of the set of concentrations
9.	Multiply the standard deviation by two to obtain the minimum detection limit.
5.3.3 Linearity
Two types of linearity will be evaluated. The first will be the linearity of the monitor for a
specific gas over a range of concentrations. The second will be the linearity of the monitor as a series
of neutral density filters are inserted into the beam path. This second evaluation of linearity is designed
to simulate a reduction in source intensity and to measure the effect this intensity reduction has on the
monitor's ability to maintain linear response.
Determining the concentration linearity of the monitor requires challenging the monitor with a
target gas at several concentration levels. At each of these concentrations, a single beam spectrum is
acquired.
The procedure for determining concentration linearity is as follows:

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 29 of 67
1.	Place the target gas cell in the optical path of the monitor
2.	Set up dilution system to provide the target gas to the gas cell by dilution of a certified gas
standard for each gas of interest
3.	Perform dilutions with high purity nitrogen
4.	Provide diluent gas or a prepared dilution of the target gas to the gas cell
5.	Choose appropriate averaging time for the monitor
6.	After at least five cell volumes of the gas have passed through the cell, acquire a single
beam spectrum
7.	Record the concentration value given by the monitor
8.	Flush cell with at least five volumes of high purity nitrogen, and again acquire a single beam
spectrum
9.	Repeat steps 4-6 with next concentration of the target gas.
The source strength linearity will be evaluated at two concentrations for each gas using three
neutral density (ND) filters placed in the beam path. These three neutral density filters will be used to
determine the monitor's ability to maintain a linear response with an attenuated source. These filters will
attenuate the source strength by approximately 10%, 25%, and 50%. The procedure for this
evaluation is identical to the steps 1 through 7 above except that one of the ND filters is placed in the
optical path.
5.3.4 Accuracy
Accuracy of the monitors relative to the gas standards will be verified by introducing known
concentrations of the target gas into the cell. The gas cell is flushed with at least five cell volumes of
nitrogen and a single beam spectrum is recorded. The target gas is then introduced into the cell and
after flushing with at least five cell volumes a single beam spectrum of the target gas is obtained. The

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 30 of 67
cell is flushed with at least five cell volumes of nitrogen and a third spectrum is recorded. The three
spectra are analyzed for the target gas using the background selected by the vendor. The concentration
of the target gas is the result of analyzing the second spectrum minus the average of the first and third
(flushed cell) spectra.
The accuracy is evaluated at concentrations cl through c4 using an integration time of 1 minute.
The accuracy is then evaluated at concentrations c2 using a longer integration time, and then again at a
concentration of c2 during the interference measurements (Activity #26 through 34). The percent
accuracy is the average value of all the measurements at the same conditions divided by the
concentration of the gas in the reference cell times 100.
5.3.5 Precision
The precision of the monitor is a quantification of its ability to make repeatable measurements
when challenged with the same gas sample. The procedure for the determination of precision is
essentially identical to the procedure for the determination of accuracy. The gas cell is flushed with at
least five volumes of nitrogen. The target gas is then introduced into the cell and after flushing with at
least five cell volumes 25 single beam spectra of the target gas are obtained. These spectra are
analyzed for the target gas. The relative standard deviation of this set of measurements is the precision
at the target gas concentration. Precision is evaluated by this procedure at two different concentrations
of each of the target gases (see Table 1). Additional precision information will be obtained from the
replicate analysis conducted in the interference test (Section 5.3.6).
5.3.6. Interferences
The effects of interfering gases will be established by supplying the reference cell with a target
gas and varying the distance between the source and detector of the monitor. The main interferences in
ambient air are H20 and C02 and if the measurements are made outdoors, changing the pathlength will

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 31 of 67
effectively change the amount of interferents in the measurement. The purpose of the interference
measurements (Activity # 26 through 34 in Table 1) is to determine the effects that the interfering gases
have on accuracy, precision, and MDL. These tests are performed using two different integration times
to determine the effect that integration time has on the monitor's ability to make measurements with
interfering gases in the light path.
The effect of the interferences will be measured by setting up the monitor outdoors, or in an
area where the light path passes through ambient air levels of H20 and C02 that are consistent with
those outdoors as measured by a relative humidity monitor and a C02 monitor (for example, in an
airplane hangar). First, the pathlength will be changed to approximately 400 meters. Then, the
reference cell will be supplied with nitrogen, and after flushing with at least five cell volumes, 5 single
beam spectra will be recorded. Next, the target gas will be introduced into the cell and after similarly
flushing the cell, 5 single beam spectra will be recorded. Finally, nitrogen will again be introduced into
the cell and 5 spectra will be recorded.
Then the pathlength will be set to the length that the vendor chooses as optimum. This is the
pathlength that would theoretically yield the best signal to noise ratio and the entire measurement
procedure will be repeated. Atmospheric concentrations of H20 and C02 will be recorded at the
beginning and the end of these measurements. The extent of interference will be calculated in terms of
sensitivity of the monitor to the interferent. The relative sensitivity will be reported.
5.4. Tuneable Diode Laser (TDL)
5.4.1.Gases
The gases and concentrations to be used for testing TDL open-path monitors are shown in
Table 4.
Table 4. Gases for Verification Testing of TDL Open-Path Monitors

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 32 of 67
Gas
Concentration
Pathlength (ppm-m)
nh3
cl
5
c2
25

c3
50

c4
100
HF
cl
25
c2
50

c3
100

c4
500
Methane
cl
5
c2
25

c3
50

c4
100

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 33 of 67
5.4.2. Minimum Detection Limit
The minimum detection limit (MDL) is to be determined for each target gas. This number
represents the lowest obtainable value for the detection of that specific gas. The MDL is calculated by
removing the target gas from the optical path of the monitor, then a series of 25 measurements are
taken using the appropriate averaging time (either 1 min. or 5 min.). The resulting values are then to be
analyzed for the target gas. Two times the standard deviation of the calculated concentrations is
defined as the minimum detection limit.
The procedure for determining the MDL is as follows:
1.	Remove the target gas from the optical path of the monitor.
2.	Choose appropriate averaging time for the monitor
3.	Acquire 25 measurements
4.	Analyze each absorption spectrum for the target gas.
5.	Calculate the standard deviation of the set of concentrations.
6.	Multiply the standard deviation by two to obtain the minimum detection limit.
5.4.3 Linearity
Two types of linearity will be evaluated. The first will be the linearity of the monitor for a
specific gas over a range of concentrations. The second will be the linearity of the monitor as a series
of neutral density filters are inserted into the beam path. This second evaluation of linearity is designed
to simulate a reduction in source intensity and to measure the effect this intensity reduction has on the
monitor's ability to maintain linear response.
Determining the concentration linearity of the monitor requires challenging the monitor with a
target gas at several concentration levels. At each of these concentrations a measurement is made.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 34 of 67
The procedure for determining concentration linearity is as follows:
1.	Place the target gas cell in the optical path of the monitor.
2.	Set up dilution system to provide the calibration gas to the gas cell by dilution of a certified
gas standard for each gas of interest
3.	Perform dilutions with high purity nitrogen
4.	Provide target gas or a prepared dilution of the target gas to the gas cell
5.	Choose appropriate averaging time for the monitor
6.	After at least five cell volumes of the gas have passed through the cell, make measurements
7.	Record the concentration value given by the monitor
8.	Flush cell with at least five volumes of high purity nitrogen, and again make a measurements
9.	Repeat steps 4-6 with next concentration.
The source strength linearity will be evaluated at two concentrations for each gas using three
neutral density (ND) filters placed in the beam path. These three neutral density filters will be used to
determine the monitor's ability to maintain a linear response with an attenuated source. These filters will
attenuate the source strength by approximately 10%, 25%, and 50%. The procedure for this
evaluation is identical to steps 1 through 7 above except that one of the ND filters is placed in the
optical path.
5.4.4 Accuracy
Accuracy of the monitors relative to the gas standards will be verified by introducing the target
gas into the cell. The gas cell is flushed with at least five cell volumes of nitrogen and a measurement is
recorded. The target gas is then introduced into the cell and after flushing with at least five cell volumes

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 35 of 67
a measurement of the target gas is obtained. The cell is flushed with at least five cell volumes of
nitrogen and a third measurement is recorded. The three measurements are analyzed for the target gas
using the background selected by the vendor. The concentration of the target gas is the result of
analyzing the second measurement minus the average of the first and third (flushed cell) measurements.
The accuracy is evaluated at concentrations cl through c4 using an integration time of 1
minute. The accuracy is then evaluated at concentrations c2 using a longer integration time, and then
again at a concentration of c2 during the interference measurements (Activity #26 through 34). The
percent accuracy is the average value of all the measurements at the same conditions divided by the
concentration of the gas in the reference cell times 100.
5.4.5	Precision
The precision of the monitor is a quantification of its ability to make repeatable. measurements
when challenged with the same gas sample. The procedure for the determination of precision is
essentially identical to the procedure for the determination of accuracy. The gas cell is flushed with at
least five volumes of nitrogen. The target gas is then introduced into the cell and after flushing with at
least five cell volumes, 25 measurements of the target gas are obtained. The relative standard deviation
of this set of concentrations is the precision at the target gas concentration. Precision is evaluated by
this procedure at two different concentrations of each of the target gases (see Table 1). Additional
precision information will be obtained from the replicate analysis conducted in the interference
measurements (Section 5.4.6)
5.4.6	Interferences
The effects of interfering gases will be established by supplying the reference cell with a target
gas and varying the distance between the source and detector of the monitor. The main interferences in
ambient air are H20 and C02 and if the measurements are made outdoors, changing the pathlength will

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 36 of 67
effectively change the amount of interferents in the measurement. The purpose of the interference
measurements (#26 through 34 in Table 1) is to determine the effects that the interfering gases have on
accuracy, precision, and MDL. These tests are performed using two different integration times to
determine the effect that integration time has on the monitor's ability to make measurements with
interfering gases in the light path.
The effect of the interferences will be measured by setting up the monitor outdoors, or in an
area where the light path passes through ambient air levels of H20 and C02 that are consistent with
those outdoors as measured by the relative humidity monitor and the C02 monitor (for example, in an
airplane hangar). First, the pathlength will be changed to approximately 400 meters. Then, the
reference cell will be supplied with nitrogen, and after flushing with at least five cell volumes, 5 single
beam spectra will be recorded. Next, the target gas will be introduced into the cell and after similarly
flushing the cell, 5 single beam spectra will be recorded. Finally, nitrogen will again be introduced into
the cell and 5 spectra will be recorded.
Then the pathlength will be set to the length that the vendor chooses as optimum. This is the
pathlength that would theoretically yield the best signal to noise ratio and the entire measurement
procedure will be repeated. Atmospheric concentrations of H20 and C02 will be recorded at the
beginning and the end of these measurements. The extent of interference will be calculated in terms of
sensitivity of the monitor to the interferent. The relative sensitivity will be reported.
5.5. Ultra Violet (UV) Open Path Monitors
5.5.1 Gases
The gases and concentrations to be used for testing UV open-path monitors are shown in
Table 5.
5.5.2 Minimum Detection Limit

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 37 of 67
The minimum detection limit (MDL) is to be determined for each target. This number
represents the lowest obtainable value for the detection of that specific gas. The MDL is calculated by
removing the target gas from the optical path of the monitor, then a series of 25 measurements are
taken using the appropriate averaging time (either 1 min. or 5 min.). The
Table 5. Gases for Verification Testing of UV Open-Path Monitors
Gas
Concentration
Pathlength (ppm-m)
nh3
cl
3
c2
6
c3
10
c4
20
NO
cl
2
c2
5
c3
10
c4
15
Benzene
cl
2
c2
3
c3
5
c4
10
resulting values are then to be analyzed for the target gas. Two times the standard deviation of the
calculated concentrations is defined as the minimum detection limit.
The procedure for determining MDL is as follows:

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 38 of 67
1.	Remove the target gas from the optical path of the monitor
2.	Choose appropriate averaging time for the monitor
3.	Acquire 25 measurements
4.	Analyze each measurement for the target gas
5.	Calculate the standard deviation of the set of measurements
6.	Multiply the standard deviation by two to obtain the minimum detection limit.
Additional MDL information will be obtained from the replicate analysis conducted in the
interference measurement (Section 5.5.6).
5.5.3 Linearity
Two types of linearity will be evaluated. The first will be the linearity of the monitor for a
specific gas over a range of concentrations. The second will be the linearity of the monitor as a series
of neutral density filters are inserted into the beam path. This second evaluation of linearity is designed
to simulate a reduction in source intensity and to measure the effect this intensity reduction has on the
monitor's ability to maintain linear response.
Determining the concentration linearity of the monitor requires challenging the monitor with a
target gas at several concentration levels. At each of these concentrations, a measurement is made.
The procedure for determining concentration linearity is as follows:
1.	Place the gas cell in the optical path of the monitor
2.	Set up dilution system to provide the calibration gas to the gas cell by dilution of a certified
gas standard for each gas of interest
3.	Perform dilutions with high purity nitrogen
4.	Provide target gas or a prepared dilution of the target gas to the gas cell

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 39 of 67
5.	Choose appropriate averaging time for the monitor
6.	After five cell volumes of the gas have passed through the cell, make a measurement
7.	Record the concentration value given by the monitor
8.	Flush cell with five volumes of high purity nitrogen, and again make a measurement
9.	Repeat with next concentration
10.	Repeat steps 4-6 with next concentration.
The source intensity linearity will be evaluated at two concentrations for each gas using three
neutral density (ND) filters placed in the beam path. These three neutral density filters will be used to
determine the monitor's ability to maintain a linear response with an attenuated source. These filters will
attenuate the source strength by approximately 10%, 25%, and 50%. The procedure for this
evaluation is identical to the steps 1 through 7 above except that one of the ND filters is placed in the
optical path.
5.5.4 Accuracy
Accuracy of the monitors relative to the gas standards will be verified by introducing the target
gas into the cell. The gas cell is flushed with at least five cell volumes of nitrogen and a measurement is
recorded. The target gas is then introduced into the cell and after flushing with at least five cell volumes
a measurement of the target gas is obtained. The cell is flushed with at least five cell volumes of
nitrogen and a third measurement is recorded. The three measurements are analyzed for the target gas
using the background selected by the vendor. The concentration of the target gas is the result of
analyzing the second measurement minus the average of the first and third (flushed cell) measurements.
The accuracy is evaluated at concentrations cl through c4 using an integration time of 1 minute.
The accuracy is then evaluated at concentrations c2 using a longer integration time, and then again at a

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 40 of 67
concentration of c2 during the interference measurements (Activity #26 through 34). The percent
accuracy is the average value of all the measurements at the same conditions divided by the
concentration of the gas in the reference cell times 100.
5.5.5	Precision
The precision of the monitor is a quantification of its ability to make repeatable. measurements
when challenged with the same gas sample. The procedure for the determination of precision is
essentially identical to the procedure for the determination of accuracy. The gas cell is flushed with at
least five volumes of nitrogen. The target gas is then introduced into the cell and after flushing with at
least five cell volumes, 25 measurements of the target gas are obtained. The relative standard deviation
of this set of concentrations is the precision at the target gas concentration. Precision is evaluated by
this procedure at two different concentrations of each of the target gases (see Table 1). Additional
precision information will be obtained from the replicate analysis conducted in the interference
measurements (Section 5.5.6)
5.5.6	Interferences
The effects of interfering gases will be established by supplying the reference cell with a target
gas and varying the distance between the source and detector (pathlength) of the monitor. The main
interferences in ambient air are 02 and 03 and if the measurements are made outdoors, changing the
pathlength will effectively change the amount of interferents in the measurement. The purpose of the
interference measurements (#26 through 34 in Tablel.) is to determine the effects that the interfering
gases have on the accuracy, precision and MDL. These tests are performed using two different
integration times to determine the effect that integration time has on the monitor's ability to make
measurements with interfering gases in the light path.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 41 of 67
The effect of the interferences will be measured by setting up the monitor outdoors, or in an
area where the ambient levels of 02 and 03 that are consistent with those outdoors as measured by the
02 and 03 monitors. First, the pathlength will be changed to approximately 400 meters. Then, the
reference cell will be supplied with nitrogen, and after flushing with at least five cell volumes, 5
measurements will be recorded. Next, the target gas will be introduced into the cell and after similarly
flushing the cell, 5 measurements will be recorded. Finally, the cell will be flushed again and 5 more
spectra will be recorded. Atmospheric concentrations of 02 and 03 will be recorded at the beginning
and the end of these measurements.
Then pathlength will be set to the length that the vendor chooses as optimum, the pathlength
that would theoretically yield the best signal to noise ratio, and the entire measurement procedure will
be repeated. The extent of interference will be calculated in terms of sensitivity of the monitor to the
interferent. The relative sensitivity will be reported.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 42 of 67
6. Site Description
Under this test/QA plan, the verification of each monitor will occur at Battelle's Columbus
facilities or at a location near the vendor's establishment. If the test is to be performed at the vendor's
location, the specific test site will be identified by the vendor and reviewed with Battelle prior to Battelle
staff traveling to the vendor's location to initiate the test.
At either location, the test site will be outside in an open field or parcel of land where a line of
sight is available that meets the maximum pathlength required (400 m). The site needs to be away from
local sources of emissions and yet easily accessible and able to be reached conveniently throughout the
test period. If the test site has limited access, the host (either Battelle or the vendor) must take
appropriate arrangements to ensure that all non-host staff have access. Sufficient lighting must be
available in the event that the test runs into the evening.
7. Materials and Equipment
7.1. Standard Gases
The standard gases diluted to produce target gas levels for the verification testing shall be NIST
traceable gases when possible. Alternatively, commercially certified gas will be used if NIST traceable
gases are not available for a particular analyte. The gases will be obtained in concentrations
appropriate for dilution to the concentrations required for the tests.
7.2. Dilution Gas
The dilution gas for the verification testing will be high purity nitrogen and will be supplied by
Battelle. The dilution gas must have the following specifications: Acid Rain CEM Zero Nitrogen or

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 43 of 67
equivalent (i.e. having the following purity specifications: total hydrocarbons, S02 and NOx <0.1 ppm,
CO and 02, <0.5 ppm, C02 <1 ppm, and water <5 ppm).

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 44 of 67
7.3. Dilution System
The dilution system used for preparation of the target gases must have mass flow capabilities
with an accuracy of approximately ħ 1 percent. The dilution system must be capable of accepting a
flow of compressed gas standard and diluting it with high purity nitrogen or air. It must be able to
perform dilution ratios from 1:1 to at least 100:1. The dilution system may be commercially available or
assembled from separate commercial components.
7.4. Temperature Sensor
The temperature sensor used to monitor the ambient air and test cell temperatures will be a
thermocouple with a commercial digital temperature readout. This sensor will be operated in
accordance with the manufacturer's instructions, and must have been calibrated against a certified
temperature measurement standard within the six months preceding the verification test.
7.5. Relative Humidity (RH) Sensor
The RH sensor used to determine the ambient air humidity will be a commercial RH/Dew Point
monitor that uses the chilled mirror principle. This sensor will be operated in accordance with the
manufacturer's instructions, which call for cleaning of the mirror and re-balancing of the optical path
when necessary, as indicated by the diagnostic display of the monitor. The manufacturer's accuracy
specification of this monitor must be approximately +/-5 percent RH.
7.6. Oxygen Sensor
A commercial electrochemical oxygen sensor will be used to measure the oxygen content of the
ambient air during interference measurements. The sensor will be operated according to the

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 45 of 67
manufacturer's directions, which call for zeroing with nitrogen and calibrating with ambient air (i.e.,
oxygen content of 20.9 percent).
7.7. Carbon Dioxide Sensor
A commercial non-dispersive infrared (NDIR) instrument will be used to monitor the level of
CO2 in ambient air during interference measurements. This sensor will be operated in accordance with
the manufacturer's instructions, and will be calibrated with a commercially prepared cylinder standard
of C02 in air.
7.8. Ozone Sensor
The sensor used to determine ozone in ambient air will be a commercial UV absorption monitor
designated by U.S. EPA as an Equivalent Method for this measurement. The UV absorption method is
preferred for this application over the Reference Method (which is based on ethylene
chemiluminescence) because the UV method is inherently calibrated, and requires no reagent gases or
calibration standards. This sensor will be operated in accordance with the manufacturer's instructions.
7.9. Monitor for NO and NH3
The concentrations of NO and NH3 prepared by the dilution system during testing will be
checked using a commercial EPA Reference chemiluminescent NO/NOx monitor, equipped with a high
temperature converter for reduction of NH3 to NO for detection. The monitor and converter will be
operated according to the manufacturer's instructions, and the conversion efficiency of the NH3
converter will be determined in the laboratory before each use in verification testing.
7.10. Carbon Monoxide Sensor

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 46 of 67
A commercial non-dispersive infrared (NDIR) instrument will be used to monitor the
level of CO in ambient air during interference measurements. This sensor will be operated in
accordance with the manufacturer's instructions, and will be calibrated with a commercially prepared
cylinder standard of CO in air.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 47 of 67
8. Quality Assurance/Quality Control
8.1. Calibration
8.1.1	Gas Dilution System
The gas dilution system will be the responsibility of Battelle. Flow controllers in this system
must be calibrated prior to the start of the verification test for each monitor by means of a soap bubble
flow meter. Corrections will be applied to the bubble meter data for temperature and water content.
8.1.2	Temperature Sensor
The thermocouple calibration will be based upon its comparison to a certified standard within
the six months preceding the test. The accuracy of the thermocouple will also be checked at least once
during verification testing, by comparison to a standard mercury-in-glass type thermometer. Agreement
within 3°C is required, or the thermocouple will be replaced. That comparison will be conducted as
part of the Performance Evaluation Audit procedure in Section 8.2.2.
8.1.3	Relative Humidity (RH) Sensor
The relative humidity sensor will be operated according to the manufacturer's directions, and
will employ the manufacturer's calibration. The accuracy of the monitor for RH will also be checked at
least once during verification testing for each monitor, by comparison to a standard wet/dry bulb
measurement. Accuracy within +/- 5 percent RH is required, or the calibration of the monitor will be
adjusted. That comparison will be conducted as part of the Performance Evaluation Audits in Section
8.2.2.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 48 of 67
8.1.4	Oxygen Sensor
The oxygen monitor will be calibrated on each day of use by zeroing with high purity
nitrogen and calibrating with ambient air (i.e., 20.9 percent 02), per the manufacturer's directions. This
procedure assures that the monitor calibration accounts for the effects of day-to-day variations in air
temperature and humidity. Also, at least once during the verification test of each monitor, the oxygen
monitor will be checked by comparison with another oxygen monitor. Agreement within 0.5 percent
oxygen is required when sampling ambient air, or the monitor will be recalibrated. That comparison will
be conducted as part of the Performance Evaluation Audits in Section 8.2.2.
8.1.5	Carbon Monoxide Sensor
The NDIR CO monitor will be calibrated before testing of each vendor's open-path system,
using a commercially prepared certified standard of CO. Also, at least once during the verification test
for each open-path monitor, the CO monitor will be challenged with an equally certified independent
calibration standard obtained from another supplier. Agreement must be within +/- 10 percent, or the
monitor will be recalibrated. That comparison will be conducted as part of the Performance Evaluation
Audits described in Section 8.2.2.
8.1.6	Carbon Dioxide Sensor
The NDIR C02 monitor will be calibrated before testing of each vendor's open-path system,
using a commercially prepared certified standard of C02 in air. Also, at least once during the
verification test for each open-path monitor, the C02 monitor will be challenged with an equally
certified independent calibration standard obtained from another supplier. Agreement must be within
+/-10 percent, or the monitor will be recalibrated. That comparison will be conducted as part of the
Performance Evaluation Audits described in Section 8.2.2.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 49 of 67
8.1.7	Ozone Sensor
The UV absorption method of ozone measurement is inherently calibrated, relying as it does on
the accurately determined absorption coefficient of ozone. As a result, routine calibration of the ozone
monitor is not needed. However, the monitor will be operated according to the manufacturer's
directions, with careful attention to the diagnostic indicators that assure proper operation of the monitor.
In addition, at least once during the verification test of each open-path monitor, the ozone monitor will
be checked in a side-by-side comparison with a different ozone monitor while sampling ambient air.
Agreement within 5 ppbv or 10 percent of reading, whichever is greater, is required. Failure to meet
this specification will result in investigation of the diagnostics of both monitors.
8.1.8	Monitor for NO and NH3
The NO monitor will be calibrated using a commercial standard of NO in nitrogen, the
concentration of which has been established by direct comparison to a Standard Reference Material of
NO in nitrogen, obtained from the NIST. A multipoint calibration will be performed before any
verification testing takes place, and a single point span check will be performed before testing of each
vendor's open-path system. If that single-point check differs from the original multipoint result by more
than 5 percent, then a new multipoint calibration will be performed. In addition, at least once during the
verification testing of each open-path monitor, using NO or NH3 as a target gas, the NO calibration will
be checked by measurement of an independent NO calibration standard obtained from an independent
supplier. That comparison will be conducted as part of the Performance Evaluation Audits in Section
8.2.2.
The conversion efficiency of the NH3 converter will be established before testing each open-

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 50 of 67
path system for which NH3 is a target compound. The efficiency test will consist of operating the NO
monitor with the NH3 converter, while sampling a constant NH, concentration in clean dilution gas. A
second NH3 converter will then be inserted into the sampling line downstream of the first, and the
conversion efficiency of the first converter will be assessed by the increase (if any) in response. This
approach requires a stable NH3 source, but does not require a certified or even known NH3
concentration. All NH? measurements will be corrected for the conversion efficiency determined in this
way.
8.2 Assessment and Audits
8.2.1	Technical Systems Audit
Battelle's Quality Manager, Ms. Sandy Anderson or her designee, will perform a technical
systems audit once during a verification test. The purpose of this technical systems audit is to ensure
that the verification test is being performed in accordance with this test/QA plan and that all QA/QC
procedures are being implemented. During this audit, Ms. Anderson will review the calibration sources
and methods used, compare actual test procedures to those specified in this plan, and review data
acquisition and handling procedures. She will also review instrument calibration records and gas
certificates of analysis.
8.2.2	Performance Evaluation Audits
Performance evaluation audits will be conducted to assess the quality of the measurements
made in this verification test. These audits address only those measurements made by Battelle in
conducting the verification test, i.e., the monitors being verified and the vendors operating these
analyzers are not the subject of the performance evaluation audits. These audits will be performed by
analyzing a standard or comparing to a reference that is independent of standards used during the

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 51 of 67
testing. These audits will be performed once during the verification test of each monitor. The audit
procedures, which are listed in Table 6, will be performed by the technical staff responsible for the
measurements being audited. Battelle's Quality Manager will be present during at least one of the
performance evaluation audits to assess the results.
The measurements (physical or chemical) will undergo the performance evaluation audit by
comparison to independent measurements or standards, as indicated in Sections 8.1.2 through 8.1.8,
and summarized in Table 6. If during the performance evaluation audit, the measurement

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 52 of 67
Table 6. Summary of Performance Evaluation Audit Procedures(a)
Measurement to be Audited
Audit Procedure
Ammonia
Compare measurements using
an independent NO standard.
Temperature
Compare to independent
temperature measurement (Hg
thermometer)
Relative humidity
Compare to independent RH
measurement (wet/dry bulb
device)
Oxygen
Compare to independent
oxygen measurement (different
analyzer)
Carbon Monoxide
Compare measurement using
an independent carbon
monoxide standard
Carbon Dioxide
Compare measurement using
an independent carbon dioxide
standard
Ozone
Compare to independent ozone
measurement (different
analyzer)
NO
Compare measurements using
an independent NO standard
Other target gases (e.g.
benzene, methane, etc.)
Comparison to results of gas
chromatographic analysis of
canister sample
HF
Comparison to results of ion
selective electrode or ion
chromatography analysis of
impinger sample

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 53 of 67
(a) Each audit procedure will be performed at least once during the
verification test.
being audited does not meet the specified performance criteria, the verification test will be stopped until
the cause of the failed audit is determined.
Table 6 indicates that performance auditing of the prepared HF and hydrocarbon
concentrations will be conducted by independent analysis of the test gas mixture supplied to the optical
cell during verification testing. For the target organic compounds (i.e., methane, benzene, ethylene,
cyclohexane, and tetrachloroethylene), this procedure will involve collecting a sample of the test gas
mixture exiting the cell using a pre-cleaned and evacuated Summa-polished sampling canister. This gas
sample will be returned to Battelle, and analyzed using EPA Method 18 when applicable for the target
hydrocarbons (methane, benzene, ethylene, and cyclohexane) using gas chromatography with flame
ionization detection (GC/FID). Calibration of the FID response will be based on a NIST propane
standard containing 9 ppm carbon. Analysis for tetrachloroethylene will be by GC with mass selective
detection (GC/MS), using a gravimetrically prepared standard of the target compound for calibration.
For HF, the performance audit will involve passing a known volume of the gas mixture exiting
the optical cell through an impinger containing deionized water. The collected HF solution will then be
analyzed at Battelle by either of two techniques: an ion selective electrode, as is the basis for EPA
Method 13B, or ion chromatography for fluoride ion. In either case calibration will be based on
fluoride solution standards prepared gravimetrically from high purity water and reagents.
For both the organics and HF, the analytical results of the performance audit samples must
indicate concentrations in the optical cell within 10 percent of the expected concentrations. If not, the
target gas source and dilution system will be assembled at Battelle, and additional samples will be
collected and analyzed to re-establish the output of the gas source and dilution system. The same
optical cell used in the verification test will be obtained from the technology vendor for use in this effort.
8.2.3 Data Quality Audit
Battelle's Quality Manager will audit at least 10 percent of the verification data acquired in the

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 54 of 67
verification test. The Quality Manager will trace the data from initial acquisition, through reduction and
statistical comparisons, and to final reporting. All calculations performed on the data undergoing audit
will be checked.
8.3 Assessment Reports
Each assessment and audit will be documented in accordance with Section 2.9.7 of the QMP
for the AMS pilot.(1) Assessment reports will include the following:
•	Identification of any adverse findings or potential problems
•	Response to adverse findings or potential problems
•	Possible recommendations for resolving problems
•	Citation of any noteworthy practices that may be of use to others
Confirmation that solutions have been implemented and are effective.
8.4 Corrective Action
The Quality Manager during the course of any assessment or audit will identify to the technical
staff performing experimental activities any immediate corrective action that should be taken. If serious
quality problems exist, the Battelle Quality Manager is authorized to stop work. Once the assessment
report has been prepared, the Verification Test Coordinator will ensure that a response is provided for
each adverse finding or potential problem, and will implement any necessary follow-up corrective
action.
9. Data Analysis and Reporting
9.1. Data Acquisition

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 55 of 67
Data acquisition in this verification test includes recording of response data from the monitors
undergoing testing, operational data such as ambient RH and temperatures, times of test activities, etc.
Data acquisition for the commercial monitors undergoing verification is primarily performed by
the vendors themselves during the test. Each monitor must have some form of data acquisition device,
such as a digital display whose readings can be recorded manually, a printout of the monitor's
response, or an electronic data recorder that stores individual monitor results. Throughout the test the
vendor will be responsible for reporting the response of the monitor to the sample gases provided.
Forms for this purpose will be provided as needed by Battelle.
Other data will be recorded in laboratory record books maintained by each Battelle staff
member involved in the testing. These records will be reviewed on a daily basis to identify and resolve
any inconsistencies.
In all cases, strict confidentiality of data from each vendor's monitor, and strict separation of
data from different monitors, will be maintained. Separate files (including manual records, printouts,
and/or electronic data files) will be kept for each monitor. At no time during verification testing will
Battelle staff engage in any comparison or discussion of test data or of different monitors.
Table 7 summarizes the types of data to be recorded, where, how often, and by whom the
recording is made, and the disposition or subsequent processing of the data. The general approach is
to record all test information immediately and in a consistent format throughout all tests. Data recorded
by the vendors is to be turned over to Battelle staff immediately upon completion of the test procedure.
Test records will then be converted to Excel spreadsheet files by a designated Battelle staff member.
Identical file formats will be used for the data from all analyzers tested, to assure uniformity of data
treatment. This process of data recording and compiling will be overseen by the Verification Test
Coordinator.
9.2. Statistical Calculations
Performance characterization is based on statistical comparisons of continuous open-path

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 56 of 67
monitor results to the known concentrations of the target gases. The following statistical procedures
will be used to make those comparisons.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 57 of 67
Table 7. Summary of Data Recording Process for the Verification Tests
Data to be Recorded
Recorded
By
Where
Recorded
When Recorded
Disposition of Data
Dates, Times, Test Events
Battelle
Data
Sheet(a)
Start of each test,
whenever testing
conditions change
Used to compile result,
manually entered into
spreadsheet as necessary
Test Parameters (temp.,
RH, etc)
Battelle
Data
Sheet(a)
Every hour during
testing
Transferred to
spreadsheet
Interference Gas
Concentrations
Battelle
Data
Sheet(a)
Before and after each
measurement of
target gas
Transferred to
spreadsheet
Target Gas
Concentrations
Battelle
Data
Sheet(a)
At specified time
during each test
Transferred to
spreadsheet
Optical Open-Path
Monitor Readings
Vendor
Data
Sheet(a)
At specified time
during each test
Transferred to
spreadsheet
(a) Sample data sheet provided in Appendix A.
9.2.1 Minimum Detection Limit
The minimum detection limit (MDL) is defined as the smallest concentration at which the
monitor's expected response exceeds the calibration curve at the background reading by two times the
standard deviation (aG) of the monitor's background reading.
MDL • *x *26
9.2.2 Linearity
Both concentration and source strength linearity will be assessed by linear regression with the
certified gas concentration as independent variable and the monitor's response as dependent variable.
Linearity will be assessed in terms of the slope, intercept, and correlation coefficient of the linear

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 58 of 67
regression.
y • •Mx • *b
where >' is the response of the monitor to a reference gas, x is the concentration of the target gas in the
optical cell, M is the slope of the linear regression curve, and b is the zero offset.
9.2.3 Accuracy
The relative accuracy (A) of the monitor with respect to the reference gas will be assessed by:
where the bars indicate the mean of the reference (R) values and monitor (T) results. This parameter
will be determined at each concentration.
9.2.4 Precision
Precision will be reported in terms of the percent relative standard deviation (RSD) of a group
of similar measurements. For a set of measurements given by Tb T2, ..., Tn, the standard deviation
(o)of these measurements is:
R-T
A =
x 100
R
1/2
where T is the average of the monitor's readings. The RSD is calculated from:

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 59 of 67
RSD =
x 100

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 60 of 67
and is a measure of the measurement uncertainty relative to the absolute value of the measurement.
This parameter will be determined at each concentration.
9.2.5 Interferences
The extent of interference will be calculated in terms of sensitivity of the monitor to the
interferent species, relative to its sensitivity to the target gas at a fixed integration time. The relative
sensitivity is calculated as the ratio of the observed response of the monitor to the actual concentration
of the interferent. For example, a monitor that reports 26 ppb of cyclohexane in air with an interference
concentration of 100 ppb of C02 may report 30 ppb of cyclohexane when the C02 concentration is
changed to 200 ppb. This would result in an interference effect of (30 ppb - 26 ppb)cydohexame/(200ppb
-100 ppb)C02 or 4 percent.
9.3. Data Review
Records generated by Battelle staff in the verification test will receive one-over-one review
within two weeks after generation, before these records are used to calculate, evaluate, or report
verification results. These records may include laboratory record books; equipment calibration
records; and data sheets used to record the monitor's response. This review will be performed by a
Battelle technical staff member involved in the verification test, but not the staff member that originally
generated the record. The review will be documented by the person performing the review by adding
his/her initials and date to a hard copy of the record being reviewed. This hard copy will then be
returned to the Battelle staff member who generated or who will be storing the record.
In addition, data calculations performed by Battelle will be spot-checked by Battelle technical
staff to ensure that calculations are performed correctly. Calculations to be checked include
determination of each monitor's precision, accuracy, minimum detection limit, and other statistical
calculations identified in Section 9.2 of this test/QA plan.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 61 of 67
9.4. Reporting
Statistical data calculations that result from each of the tests described above will be conducted
separately for each optical open-path monitor. Separate verification reports will then be prepared,
each addressing the monitor provided by one commercial vendor. For each parameter evaluated in this
verification test, the verification report will present the measurement data, as well as the results of the
statistical evaluation of those data.
The verification report will briefly describe the ETV program and the AMS pilot, and will
describe the procedures used in the verification test, but will include specific requirements or departures
from procedure necessitated in testing of the individual monitor in question. These sections will be
common to each verification report resulting from this verification test. The results of the verification
test will then be stated quantitatively, without comparison to any other monitor tested, or any comment
on the acceptability of the monitor's performance. The preparation of draft verification reports, the
review of reports by vendors and others, the revision of the reports, final approval, and the distribution
of the reports, will be conducted as stated in the Generic Verification Protocol for the Advanced
Monitoring Systems Pilot.3 Preparation, approval, and use of verification statements summarizing the
results of this test will also be subject to the requirements of that same Protocol.
10. Health and Safety
10.1. General
The health and safety officer of the test facility, whether Battelle or one of the monitor vendors,
will review the necessary health and safety requirements and guidelines for the facility with Battelle and
vendor staff before the verification test begins. Battelle staff involved in this verification test will operate
under these established requirements and guidelines as well as under appropriate procedures covered in
the Battelle Safety Manual. Specifically, the use of personal protective equipment, as defined in

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 62 of 67
procedure SIH-PP-01, will be used, and the chemical safety
protocols set forth in SIH-PP-05 will be followed. It is expected that while on Battelle's site, all
vendor representatives will operate according to the Battelle site requirements.
10.2. Potential Hazards
Vendor staff will only be operating their open-path monitors during the verification test. They
are not responsible for, nor permitted to, generate dilution gases, or perform any other verification
activities identified in this test/QA plan. Operation of the open-path monitors does not pose any known
chemical, fire, mechanical, electrical, noise, or other potential hazard.
10.3. Training
All vendor staff will be given a safety briefing prior to their installation and operation of their
monitors in Battelle laboratories. This briefing will include a description of emergency operating
procedures (i.e., in case of fire, tornado, bomb, laboratory accident) and identification and location and
operation of safety equipment (e.g., fire alarms, fire extinguishers, eye washes, exits). Similar instruction
will be provided by the vendor to all Battelle staff members traveling to the vendor's site.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 63 of 67
11. Definitions
Accuracy - The degree of agreement between the response of the optical open-path monitor and
actual gas concentration.
Dilution System - An instrument or apparatus equipped with mass flow controllers, capable of flow
control to ħ1 percent accuracy, and used for dilution of the target gas to concentrations suitable for the
testing of the monitors.
Monitor - System provided by vendor, consisting of a radiation source and detector, used to measure
atmospheric pollutants.
Minimum Detection Limit - The concentration at which the response of the optical open-path
monitor equals two times the standard deviation of the noise level at the monitor background.
Linearity - The linear proportional relationship expected between analyte concentration and monitor
response over the full measuring range of the monitor.
Precision - The degree of mutual agreement among successive readings of the same sample gas.
Neutral Density Filter - An optical filter that attenuates an incident beam of radiation without
changing its spectral distribution, that is, it has a constant transmittance over a wide spectral range.
Interference - The response of the monitor to a constituent of the sample gas other than the target gas.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 64 of 67
Pathlength - The linear distance over which the radiation from the optical open-path monitor travels
between the source and detector.
Target gas - The gas for which the monitor is making its measurement.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 65 of 67
12. References
1.	Quality Management Plan (QMP) for the ETV Advanced Monitoring Systems Pilot, U. S. EPA
Environmental Technology Verification Program, prepared by Battelle, Columbus, Ohio,
September 1998.
2.	Compendium Method TO-16 Long-Path Open-Path Fourier Transform Infrared Monitoring of
Atmospheric Gases, EPA-625/R-96/010b, U.S. Environmental Protection Agency, Cincinnati,
Ohio, January 1997.
3.	Generic Verification Protocol for the Advanced Monitoring Systems Pilot, U. S. EPA
Environmental Technology Verification Program, prepared by Battelle, Columbus, Ohio, October
1998.

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 66 of 67
APPENDIX A
EXAMPLE DATA SHEET

-------
Optical Open-Path Monitors
Test/QA Plan
Revision 1
Date: 10/28/99
Page: 67 of 67
ETV Advanced Monitoring Systems Pilot
Verification of Optical Open-Path Monitor
Vendor 	
Instrument
iinnnlc GflS!
Hate:
Operator:






RĞ
'icwed hvr





/leasurement #












"ell Temp (F)












Ambient 09 Concentrations (ppb)












Ambient C09 Concentrations (ppb)












Ambient RH (%)












Ambient 0, Concentrations (ppb)












Ambient Temp (F)












ntegration Time












'athlength












Concentration in Cell












"ell Length


























-------