US Environmental Protection Agency
Office of Pesticide Programs
Office of Pesticide Programs
Microbiology Laboratory
Environmental Science Center, Ft. Meade, MD
Standard Operating Procedure for Use and Maintenance
of Biological Safety Cabinets
SOP Number: QC-06-05
Date Revised: 10-25-11

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 1 of 10
EPA/OPP MICROBIOLOGY LABORATORY
ESC, Ft. Meade, MD
Standard Operating Procedure
for
Use and Maintenance of Biological Safety Cabinets
SOP Number: QC-06-05
Date Revised: 10-25-11
Initiated By:		Date:	/	/_
Print Name:	
Technical Review: 	Date: / /
Print Name:
Technical Staff
QA Review:		Date: / /
Print Name:	
QA Officer
Approved By:		Date: / /
Print Name:	
Branch Chief
Effective Date: 	/	/	
Controlled Copy No.:	
Withdrawn By:
Date: / /

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 2 of 10
Contents	Page Number
1.0 SCOPE AM) APPLICATION	3
2.0 DEFINITIONS	3
3.0 HEALTH AND SAFETY	3
4.0 CAUTIONS	3
5.0 INTERFERENCES	3
6.0 PERSONNEL QUALIFICATIONS	4
7.0 SPECIAL APPARATUS AND MATERIALS	4
8.0 INSTRUMENT OR METHOD CALIBRATION	5
9.0 SAMPLE HANDLING AND STORAGE	5
10.0 PROCEDURE AM) ANALYSIS	5
11.0 DATA ANALYSIS/CALCULATIONS	7
12.0 DATA MANAGEMENT/RECORDS MANAGEMENT	7
13.0 QUALITY CONTROL	7
14.0 NONCONFORMANCE AND CORRECTIVE ACTION	8
15.0 REFERENCES	8
16.0 FORMS AM) DATA SHEETS	9

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 3 of 10
1.0 SCOPE AND APPLICATION:
1.1 The purpose of this SOP is to describe the procedure used to monitor the use and
maintenance of the Biological Safety Cabinets (BSC). Performing maintenance
and repairs on the BSCs and maintaining annual certification of the BSCs are the
responsibilities of the Facility Manager or his designees (e.g., operations,
maintenance and repair contractors).
2.0 DEFINITIONS:
2.1	BSC = biological safety cabinet
2.2	FPM = feet per minute
2.3	CFM = cubic feet per minute
2.4	LFPM = linear feet per minute
2.5	HEPA = high efficiency particulate air
3.0 HEALTH AND SAFETY:
3.1	Laboratory workers are required to read and be familiar with SOP MB-01, Lab
Biosafety, prior to working with microorganisms in the BSC. Documentation of
training and familiarization with SOP MB-01 can be found in the training file for
each employee.
3.2	All work involving manipulation of cultures (SOP MB-01, Lab Biosafety) of
microorganisms, regardless of biosafety level, is performed in the BSC.
4.0 CAUTIONS:
4.1	Failure to clean the ultraviolet lamps in the BSC will reduce the lamps'
effectiveness. Periodically clean the ultraviolet lamps in the BSC with a lint-free
cloth dampened with alcohol.
4.2	Quarterly cleaning of the area under the BSC work surface and grill is required.
5.0 INTERFERENCES:
5.1 The BSC must be cleaned and appropriately certified when in use.

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 4 of 10
5.2 Laboratory staff performing work in the BSC must wear a long-sleeved lab coat
and are encouraged to wear gloves even when not manipulating culture. This
minimizes the shedding of skin flora into the work area and concurrently protects
the hands and arms from viable agent contamination (see ref. 15.2, p. 41).
6.0 PERSONNEL QUALIFICATIONS:
6.1 Personnel are required to be knowledgeable of the procedures in this SOP.
Documentation of training and familiarization with this SOP can be found in the
training file for each employee.
7.0 SPECIAL APPARATUS AND MATERIALS:
7.1	BSC #1 NuAire Model No. NU-430-600 Serial No. 78536AFU located in room
B204
7.2	BSC #2 NuAire Model No. NU-430-600 Serial No. 78538AFU located in room
B204
7.3	BSC #3 NuAire Model No. NU-430-600 Serial No. 78537AFU located in room
B204
7.4	BSC #4 NuAire Model No. NU-430-600 Serial No. 78539AFU located in room
B205
7.5	BSC #5 NuAire Model No. NU-430-600 Serial No. 76788AFP located in room
B207
7.6	BSC #6 NuAire Model No. NU-430-600 Serial No. 76786AFP located in room
B207
7.7	BSC #7 NuAire Model No. NU-430-600 Serial No. 85756AGY located in room
B202
7.8	BSC #8 NuAire Model No. NU-430-600 Serial No. 85755AGY located in room
B202
7.9	BSC #11 NuAire Model No NU 430-600, Serial No. 90496042204 located in
room B201
7.10	BSC # 12 NuAire Model No. NU 430-600, Serial No. 90474042104 located in
room D122

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 5 of 10
8.0 INSTRUMENT OR METHOD CALIBRATION:
8.1 Work zone air downflow is uniform and held to an average velocity of 60 LFPM
(0.30 m/s) measured in the plane of the bottom edge of the viewing window. The
BSC will alarm if the average velocity falls below 54 LFPM or above 64 LFPM.
The significance of this airflow control is that the work zone air is flowing down
like a huge piston carrying contaminants rapidly out of the work zone with 24 air
changes per minute. The work zone airflow is ducted such that the quantity of air
leaving through the rear perforated area is half of the downflow air quantity. The
remainder, that is the other half of the downflow air, and the air entering through
the work access opening flow into the front perforated area. Room air flowing
into the 8 inch (203 mm) work access opening forms the air barrier protecting
personnel at 105 ±5 FPM (0.53 ±0.25 m/s). The exhaust volume is displayed on
the front panel. The BSC alarms if the exhaust volume falls below 980 CFM or
rises to above 1250 CFM.
9.0 SAMPLE HANDLING AND STORAGE: Not applicable
10.0 PROCEDURE AND ANALYSIS:
10.1	Turn on the blowers, lights, and outlets, and allow to operate for a minimum of 15
minutes before aseptic manipulations are begun in the BSC.
10.2	Disinfect the interior surface of the BSC work space thoroughly by
spraying/wetting with an appropriate amount of an EPA registered hospital
disinfectant and allow to stand for the label-specified contact time before wiping.
10.2.1 Halogenated materials (e.g., bleach) are not recommended for use on
stainless steel surfaces of the BSC.
10.3	Place the apparatus and materials into the cabinet. Care must be exercised that no
items be placed over the front intake grills. The work area should be arranged so
that clean and dirty (used) materials are well separated. Passage of contaminated
materials over uninoculated cultures or clean glassware should be avoided and
transfer of viable materials should be performed as deeply into the cabinet (away
from open face) as possible.
10.4	Additional purging of the work space without any user activity should be allowed
for 2-3 minutes after materials and apparatus have been placed in it. This helps
the area get rid of all contamination that may have been introduced with the items.
10.5	Record the Downflow (FPM) and Exhaust (CFM) rates on the BSC Monitoring

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 6 of 10
Record Form (see 16.1) immediately prior to use.
10.6	Work can now be performed.
10.7	Applying a Bunsen burner flame to tube and flask necks is a conventional method
of minimizing airborne contamination when the tubes or flasks are opened.
Because the laminar flow BSCs create an environment that has little to no
airborne particulates, it is not necessary to flame tube and flask necks. Gas
burners contribute to heat build-up in the cabinet and disrupt the laminar air flow.
If a procedure requires the use of a flame (e.g., to flame sterilize hooks, etc.), use
a burner with a pilot light so that the flame can be ignited only when required. Do
not use constant flame gas burners. Gas burners should only be used from the
center of the work surface to the right rear, where resulting air turbulence will
have a minimal effect. Do not use gas burners on the left of the work surface due
to its influence on the electronic airflow control system (see ref. 15.2, p. 40).
10.8	Once work is completed, allow the BSC to run for a 2-3 minute period without
activity to purge the unit. Remove all materials, cultures, and equipment and
decontaminate BSC as in 10.2.
10.8.1	As an additional step, at the conclusion of activities involving bacteria in
spore form (e.g., Bacillus subtilis spore suspensions, spore strips, etc.) turn
on the ultraviolet light and leave it on overnight.
10.8.2	Refer to the laboratory's Biosafety Plan for Bacillus anthracis for BSC
clean up procedures following work with select agents.
10.9	A careful check of grills and diffuser grids should be made for spilled or splashed
nutrients which may support microbial growth and result in the liberation of
contaminants in the protected work environment.
10.10	A permanent paper catch is installed behind the rear divider panel of the work
zone. This area forms the return air path to the motor/blower. If the air flow is
blocked, it could seriously affect the performance of the cabinet. The paper catch
should be checked and cleaned no less than weekly; daily if procedures require
the use of paper products. Any paper removed must be properly disposed of as
contaminated hazardous waste.
10.11	Turn off lights and outlets. Turn off blowers as per facility recommendations. Do
not use cabinet as a repository for excess lab equipment during periods of non-
operation.

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 7 of 10
10.12	Record maintenance on the BSC Monitoring Record Form (see 16.1).
10.13	Quarterly Maintenance:
10.13.1	A permanent paper catch is installed behind the rear divider panel of the
work zone. This area forms the return air path to the motor/blower. If
the air flow is blocked, it could seriously affect the performance of the
cabinet. The paper catch should be checked and cleaned no less than
weekly; daily if procedures require the use of paper products. Any paper
removed must be properly disposed of as contaminated hazardous waste.
10.13.2	All BSCs should be thoroughly cleaned and disinfected at least
quarterly.
10.13.3	Remove panel and grill from the BSC and set aside.
10.13.4	Disinfect all surfaces of the BSC thoroughly by spraying/wetting with an
appropriate amount of a hospital disinfectant and allow to stand for the
label-specified contact time before wiping.
10.13.5	Disinfect panel and grill thoroughly by spraying/wetting with an
appropriate amount of a hospital disinfectant and allow to stand for the
label-specified contact time before wiping.
10.13.6	Replace the panel and grill, and properly secure.
10.13.7	Record quarterly maintenance on the BSC Monitoring Record Form (see
16.1).
11.0 DATA ANALYSIS/CALCULATIONS: None
12.0 DATA MANAGEMENT/RECORDS MANAGEMENT:
12.1 Data will be recorded promptly, legibly, and in indelible ink on the BSC
Monitoring Record Form (see 16.1). Completed forms are archived in notebooks
kept in secured file cabinets in file room D217. Only authorized personnel have
access to the secured files. Archived data is subject to OPP's official retention
schedule contained in SOP ADM-03 (Records and Archives).
13.0
QUALITY CONTROL:

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 8 of 10
13.1	Certification:
13.1.1 The BSCs are recertified annually by a qualified technician. A series of
tests are performed on the BSC during recertification including the HEP A
filter leak test and a measurement of the downflow blower speed
(designated "Blower" or "Blower Speed" on the certificate and report).
13.2	Scheduling for recertification:
13.2.1 Scheduling recertification is the responsibility of the ESC Facility
Manager or a designee. Certificates are maintained by the Facility
Manager. The OPP Microbiology Laboratory requests a copy of the Test
and Certification Reports each year for review.
13.3	HEPA filter usage and replacement:
13.3.1	Under normal usage and barring an accident (puncture), the HEPA filters
on each BSC do not need replacement until the exhaust volume or the
access flow velocity cannot be maintained. Making this determination and
scheduling HEPA filter replacement is the responsibility of the Facility
Manager or his designees.
13.3.2	The Facility Manager takes into consideration the downflow blower speed
(if >90%) and other indicators (e.g., age - 8 to 10 years) to determine the
need for replacement of the HEPA filter.
13.3.3	Once the HEPA filter is replaced, the BSC is recertified and the
certification is annotated by a Performance Certification sticker.
14.0 NONCONFORMANCE AND CORRECTIVE ACTION:
14.1 When laboratory staff believe that service is necessary for the BSCs (e.g., alarms
are sounding, the BSC fails to advance past the warm-up mode, the BSC is not
functioning as it usually does, etc.), staff may notify the Facilities Helpline at
(410) 305-4357. The Facility Manager or his designees investigates and makes a
service call, if necessary. Laboratory staff notes the event on the BSC Monitoring
Record Form (16.1) and documents any corrective action taken.
15.0 REFERENCES:
15.1 Bordner, R.H., Winter, J.A. &, Scarpino, P.V., eds. 1978. Microbiological

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 9 of 10
Methods for Monitoring the Environment, Water and Wastes. EPA 600/8-78-
017, Environmental Monitoring & Support Lab., U.S. Environmental Protection
Agency, Cincinnati, Ohio.
15.2 NuAire Labguard Class II, Type B2 Laminar Flow Biological Safety Cabinet
Operation and Maintenance Manual.
16.0 FORMS AND DATA SHEETS:
16.1 BSC Monitoring Record Form

-------
SOP No. QC-06-05
Date Revised 10-25-11
Page 10 of 10
16.1 BSC Monitoring Record Form
OPP Microbiology Laboratory

Date
Init.
BSC#
Downflow (FPM)
Exhaust (CFM)
Use/Comments/Corrective Action*




















































































• Record the actions performed in the BSC whether it be transfers or Use Dilution testing, etc. Also record in this space any alarms (high flow or low flow) that occur during use and the
Exhaust (CFM) that triggered the alarm. Document actions taken when BSC fails to conform to manufacturers specifications, e.g., alarm goes off, call for service, etc.

-------