METHOD 3535A
SOLID-PHASE EXTRACTION (SPE^
1.0 SCOPE AND APPLICATION
1.1 This method describes a procedure for isolating target organic analytes from aqueous
samples using solid-phase extraction (SPE) media. The method describes conditions for extracting
a variety of organic compounds from aqueous matrices that include: groundwater, wastewater, and
TCLP leachates. The method describes the use of disk extraction media for eight groups of
analytes and the use of cartridge extraction media for one group of analytes. Other solid-phase
extraction media may be employed as described in see Sec. 4.0. The extraction procedures are
specific to the analytes of interest and vary by group of analytes and type of extraction media. The
groups of analytes that have been evaluated thus far are listed below, along with the types of media
that have been evaluated, and the determinative methods in which the corresponding performance
data can be found.
Analyte	Extraction	Determinative
Group	Media Type	Method
Phthalate esters	Disks	8061
Organochlorine pesticides	Disks	8081
Polychlorinated biphenyls (PCBs)	Disks	8082
Organophosphorus pesticides	Disks	8141
Nitroaromatics and nitramines	Disks and Cartridges	8330
TCLP leachates containing organochlorine	Disks	8081
pesticides
TCLP leachates containing semivolatiles	Disks	8270
TCLP leachates containing phenoxyacid herbicides	Disks	8321
1.2	The technique may also be applicable to other semivolatile or extractable compounds.
It may also be used for the extraction of additional target analytes or may employ other solid-phase
media, provided that the analyst demonstrates adequate performance (e.g., recovery of 70 -130%,
or project-specific recovery criteria) using spiked sample matrices and an appropriate determinative
method of the type included in Chapter Four (Sec. 4.3). The use of organic-free reagent water
alone is not considered sufficient for conducting such performance studies, and must be supported
by data from actual sample matrices.
1.3	This method also provides procedures for concentrating extracts and for solvent
exchange.
1.4	Solid-phase extraction is called liquid-solid extraction in some methods associated
with the Safe Drinking Water Act.
CD-ROM
3535A - 1
Revision 1
January 1998

-------
1.5 This method is restricted to use by or under the supervision of trained analysts. Each
analyst must demonstrate the ability to generate acceptable results with this method.
2.0 SUMMARY OF METHOD
2.1	Sample preparation procedures vary by analyte group. Extraction of some groups
requires that the pH of the sample be adjusted to a specified value prior to extraction (see Sec.
7.2). Other groups do not require a pH adjustment.
2.2	Following any necessary pH adjustment, a measured volume of sample is extracted
by passing it through the solid-phase extraction medium (disks or cartridges), which is held in an
extraction device designed for vacuum filtration of the sample.
2.3	Target analytes are eluted from the solid-phase media using an appropriate solvent
(see Sees. 7.8 and 7.9) which is collected in a receiving vessel. The resulting solvent extract is
dried using sodium sulfate and concentrated, as needed.
2.4	As necessary for the specific analysis, the concentrated extract may be exchanged
into a solvent compatible extract with subsequent cleanup procedures (Chapter Four, Sec. 4.2) or
determinative procedures (Chapter Four, Sec. 4.3) for the measurement of the target analytes.
3.0 INTERFERENCES
3.1	Refer to Method 3500.
3.2	The decomposition of some analytes has been demonstrated under basic extraction
conditions. Organochlorine pesticides may dechlorinate and phthalate esters may hydrolyze. The
rates of these reactions increase with increasing pH and reaction times.
3.3	Bonded-phase silica (e.g., C18) will hydrolyze on prolonged exposure to aqueous
samples with pH less than 2 or greater than 9. Hydrolysis will increase at the extremes of this pH
range and with longer contact times. Hydrolysis may reduce extraction efficiency or cause baseline
irregularities. Styrene divinylbenzene (SDB) extraction disks should be considered when hydrolysis
is a problem.
3.4	Phthalates are a ubiquitous laboratory contaminant. All glass extraction apparatus
should be used for this method because phthalates are used as release agents when molding rigid
plastic (e.g., PVC) and as plasticizers for flexible tubing. A method blank, as described in Chapter
One, should be analyzed, demonstrating that there is no phthalate contamination of the sodium
sulfate or other reagents listed in this method.
3.5	Sample particulates may clog the solid-phase media and result in extremely slow
sample extractions. Use of an appropriate filter aid will result in shorter extractions without loss of
method performance if clogging is a problem. Even when a filter aid is employed, this method may
not be appropriate for aqueous samples with high levels of suspended solids (>1%), as the
extraction efficiency may not be sufficient, given the small volumes of solvents employed and the
short contact time.
CD-ROM
3535A - 2
Revision 1
January 1998

-------
4.0 APPARATUS AND MATERIALS
The apparatus and materials described here are based on data provided to EPA for the
extraction of eight groups of analytes using disk-type materials and for the extraction of one group
of analytes using cartridge-type materials. Other solid-phase extraction media configurations may
be employed, provided that the laboratory demonstrates adequate performance for the analytes
of interest. The use of other SPE configurations will require modifications to the procedures
described in Sec. 7.0. Consult the manufacturer's instructions regarding such modifications.
4.1	Solid-phase disk extraction system - Empore™ manifold that holds three 90-mm filter
standard apparatus or six 47-mm standard filter apparatus, or equivalent. Other manual,
automatic, or robotic sample preparation systems designed for solid-phase media may be utilized
for this method if adequate performance is achieved and all quality control requirements are
satisfied.
4.1.1	Manifold station - (Fisher Scientific 14-378-1B [3-place], 14-378-1A [6-place],
or equivalent).
4.1.2	Standard filter apparatus - (Fisher Scientific 14-378-2A [47-mm], 14-378-2B
[90-mm], or equivalent), consisting of a sample reservoir, clamp, fritted disk and filtration
head with drip tip.
4.1.3	Collection tube- 60-mL. The collection tube should be ofappropriate ID and
length so that the drip tip of the standard filter apparatus can be positioned well into the neck
of the tube to prevent splattering.
4.1.4	Filter flask - 2-L with a ground-glass receiver joint (optional). May be used
to carry out individual disk extractions with the standard filter apparatus and collection vial in
an all-glass system.
4.2	Solid-phase cartridge extraction system - Visiprep solid-phase extraction manifold
(Supelco) or equivalent system suitable for use with the extraction cartridges (see Sec. 4.4).
Consult the manufacturer's recommendations for the associated glassware and hardware
necessary to perform sample extractions.
4.3	Solid-phase extraction disks - Empore™, 47-mm, 90-mm, or equivalent. Disks are
available in 47-mm and 90-mm diameters, composed of a variety of solid-phase materials. Other
solid phases may be employed, provided that adequate performance is demonstrated for the
analytes of interest. Guidance for selecting the specific disk is provided in Table 1.
4.3.1	C18 disks - Empore™ disks, 47-mm diameter (3M product number 98-0503-
0015-5), 90-mm diameter (3M product number 98-0503-0019-7), or equivalent.
4.3.2	C18 fast flow disks - Empore™ disks, 47-mm diameter (3M product number
98-0503-0138-5), 90-mm diameter (3M product number 98-0503-0136-9), or equivalent.
These disks may be a better choice for samples that are difficult to filter even with the use of
a filter aid.
CD-ROM
3535A - 3
Revision 1
January 1998

-------
4.3.3	Styrene divinylbenzene (SDB-XC) disks - Empore™ disks, 47-mm diameter
(3M product number 98-0503-0067-6), 90-mm diameter (3M product number 98-0503-0068-
4), or equivalent.
4.3.4	Styrene divinylbenzene reversed-phase sulfonated (SDB-RPS) disks -
Empore™ disks, 47-mm diameter (3M product number 98-0503-0110-4), 90-mm diameter
(3M product number 98-0503-0111-2), or equivalent.
4.4	Solid-phase extraction cartridges - Porapak® R SPE device, Waters Corporation, or
equivalent. Other solid phases may be employed, provided that adequate performance is
demonstrated for the analytes of interest.
4.5	Filtration aid (optional)
4.5.1	Filter Aid 400- (Fisher Scientific 14-378-3, or equivalent).
4.5.2	In-situ glass micro-fiber prefilter - (Whatman GMF 150, 1-|jm pore size, or
equivalent).
4.6	Drying column - 22-mm ID glass chromatographic column with a PTFE stopcock
(Kontes K-420530-0242, or equivalent).
NOTE: Fritted glass discs used to retain sodium sulfate in some columns are difficult to
decontaminate after contact with highly contaminated or viscous extracts. Columns
suitable for this method use a small pad of glass wool to retain the drying agent.
4.7	Kuderna-Danish (K-D) apparatus
4.7.1	Concentrator tube - 10-mL, graduated. A ground-glass stopper is used to
prevent evaporation of extracts during short-term storage.
4.7.2	Evaporation flask - 500-mL, or other size appropriate for the volumes of
solvents to be concentrated. Attach to concentrator tube using springs or clamps.
4.7.3	Three-ball macro-Snyder column.
4.7.4	Two-ball micro-Snyder column (optional).
4.7.5	Springs - 1/4-inch.
4.8	Solvent Vapor Recovery System - Kontes 545000-1006 or K-547300-0000, Ace Glass
6614-30, or equivalent.
NOTE: The glassware in Sec. 4.6 is recommended for the purpose of solvent recovery during the
concentration procedures (Sees. 7.10 and 7.11) requiring the use of Kuderna-Danish
evaporative concentrators. Incorporation of this apparatus may be required by State or
local municipality regulations that govern air emissions of volatile organics. EPA
recommends the incorporation of this type of reclamation system as a method to
implement an emissions reduction program. Solvent recovery is a means to conform with
waste minimization and pollution prevention initiatives.
CD-ROM
3535A - 4
Revision 1
January 1998

-------
4.9	Boiling chips - Solvent extracted, approximately 10/40 mesh (silicon carbide, or
equivalent).
4.10	Water bath - Heated, with concentric ring cover, capable of temperature control to
within ± 5°C. The bath should be used in a hood.
4.11	Nitrogen evaporation apparatus (optional) - N-Evap, 12-or24-position (Organomation
Model 112, or equivalent).
4.12	Vials, glass - Sizes as appropriate, e.g., 2-mL or 10-mL, with PTFE-lined screw caps
or crimp tops for storage of extracts.
4.13	pH indicator paper - Wide pH range.
4.14	Vacuum system - Capable of maintaining a vacuum of approximately 66 cm (26
inches) of mercury.
4.15	Graduated cylinders - Sizes as appropriate.
4.16	Pipets - disposable.
5.0 REAGENTS
5.1	Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is
intended that all reagents shall conform to the specifications of the Committee on Analytical
Reagents of the American Chemical Society, where such specifications are available. Other
grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to
permit its use without decreasing the accuracy of the determination. Reagents should be stored
in glass to prevent the leaching of contaminants from plastic containers.
5.2	Organic-free reagent water - All references to water in this method refer to organic-
free reagent water, as defined in Chapter One.
5.3	Sodium sulfate (granular, anhydrous), Na2S04- Purify by heating at400°C for4 hours
in a shallow tray, or by precleaning the sodium sulfate with methylene chloride.
5.4	Solutions for adjusting the pH of samples before extraction.
5.4.1	Sulfuric acid solution (1:1 v/v), H2S04 - Slowly add 50 ml_ of concentrated
H2S04 (sp. gr. 1.84) to 50 ml_ of organic-free reagent water.
5.4.2	Sodium hydroxide solution (10N), NaOH - Dissolve 40 g NaOH in organic-
free reagent water and dilute to 100 ml_.
5.5	Extraction, washing, and exchange solvents - At a minimum, all solvents must be
pesticide quality or equivalent.
5.5.1 Methylene chloride, CH2CI2.
CD-ROM
3535A - 5
Revision 1
January 1998

-------
5.5.2
Hexane, C6H14.
5.5.3
Ethyl acetate, CH3C(OH)OCH2CH3.
5.5.4
Acetonitrile, CH3CN.
5.5.5
Methanol, CH3OH.
5.5.6
Acetone, (CH3)2CO.
5.5.7
Methyl-te/f-butyl ether (MTBE), C5H120
6.0	SAMPLE COLLECTION, PRESERVATION, AND HANDLING
See the introductory material to Chapter Four, Organic Analytes, Sec. 4.1, Method 3500, Sec.
7.1	of this method, and the specific determinative methods to be employed.
7.0 PROCEDURE
The procedures for solid-phase extraction are very similar for most organic analytes.
Therefore, this section describes procedures for sample preparation, pH adjustment, preparation
of the extraction apparatus, and extract concentration that apply to all target analytes. The
procedures for disk washing, disk conditioning, sample extraction, and sample elution vary among
the groups of analytes.
7.1 Sample preparation
Most of the specific procedures described in this method were developed for a nominal
sample size of 1 L, as this sample size is usually employed for other extraction methods such as
separatory funnel or continuous liquid-liquid extraction. This method also may be employed with
smaller samples when overall analytical sensitivity is not a concern or when high levels of the target
analytes are anticipated. However, such samples are best collected in a container of appropriate
size. The extraction of aqueous samples presents several challenges that must be considered
during sample preparation. First, the analytes of interest are often associated with the particulate
matter in the sample and sample preparation procedures must ensure that any particulates in the
original sample are included in the sample aliquot that is extracted. Secondly, the majority of the
organic analytes are hydrophobic and may preferentially adhere to the surfaces of the sample
container. For this reason, most extraction methods have traditionally specified that once the
sample has been transferred to the extraction apparatus, the sample container be rinsed with
solvent which is added to the apparatus. As a result, it is generally not appropriate to extract only
part of the sample from a sample container, e.g., 250 mL from a 1-L sample bottle.
The appropriate sample volume may vary with the intended use of the results and, in general,
is the volume necessary to provide the analytical sensitivity necessary to meet the objectives of the
project (see Chapter Two). Under ideal conditions, the sample should be collected by completely
filling the container. The sample should generally be collected without additional volume and with
little or no headspace. Thus, a 1-L sample is collected in a 1-L container, a 250-mL sample is
collected in a 250-mL container, etc.
CD-ROM
3535A - 6
Revision 1
January 1998

-------
Any surrogates and matrix spiking compounds (if applicable) are added to the sample in the
original container. The container is then recapped and shaken to mix the spiked analytes into the
sample. The extraction of some groups of analytes also requires that the pH of the sample be
adjusted to a specified value (see Table 1). When pH adjustment is necessary, it should be
performed after the surrogates and matrix spiking compounds (if applicable) have been added and
mixed with the sample. Otherwise, the recoveries of these compounds will have little relevance to
those of the target analytes in the sample.
If this approach is not possible, then a sample aliquot may be transferred to a graduated
cylinder and spiked. However, in such instances, the analyst must take great care to mix the
sample well, by shaking, to ensure a homogeneous distribution of the particulate matter and must
record the fact that the container was not rinsed.
NOTE: This method may not be appropriate for aqueous samples with greater than 1% solids,
as such samples can be difficult to filter and the extraction efficiency may be reduced as
a result of the small volumes of solvents employed and the short contact time. If the
particulate load significantly slows or prevents filtration, it may be more appropriate to
employ an alternative extraction procedure.
7.1.1	Mark the level of the sample on the outside of the sample container for later
determination of the sample volume used. Shake the container for several minutes, with the
cap tightly sealed, to ensure that any particulate matter is evenly distributed throughout the
sample.
7.1.2	Prepare a method blank from a 1-L volume of organic-free reagent water,
or a volume similar to that of the samples (e.g., a 250-mL blank should be used when the
sample size is 250 ml_, etc.). The blank may be prepared in a graduated cylinder, beaker,
or other suitable container. The frequency of method blank preparation is described in
Chapter One.
7.1.3	Add any surrogate standards listed in the determinative method to the
samples in their original containers and to the blank. For disk extractions, also add 5.0 ml_
of methanol to each sample in the original container. All samples, blanks, and QC samples
should receive the same amount of methanol. (This step is not necessary for the cartridge
extraction of nitroaromatics and nitramines.) Shake the samples to mix the surrogates and
allow the sample to stand for at least several minutes. This will permit the surrogates to
dissolve in the sample and will also allow the particulate matter to settle after spiking, which
will speed the filtration process somewhat.
7.1.4	Prepare matrix spikes by adding listed matrix spike standards to
representative sample replicates in their original containers. The frequency with which matrix
spikes are prepared and analyzed is described in Chapter One or as part of the determinative
method. Mix the matrix spike samples as described in Sec.7.1.3 and allow to stand.
7.1.5	If cleanup procedures are to be employed that result in the loss of extract,
adjust the amount of surrogate and spiking cocktail(s) accordingly. In the case of Method
3640, Gel Permeation Cleanup, double the amount of standards to compensate for the loss
of one half of the extract concentrate when loading the GPC column.
CD-ROM
3535A - 7
Revision 1
January 1998

-------
7.2 pH adjustment
Check the pH of the sample with wide-range pH paper and, if necessary, adjust the pH to the
range listed below. If pH adjustment is required, this step should be performed in the original
sample container to ensure that analytes are not lost in precipitates or flocculated material. Any
adjustment of the sample pH should take place after the surrogates and matrix spiking compounds
are added, so that they are affected by the pH in the same manner as the target analytes.
NOTE: The efficiency of solid-phase extraction of acid herbicide compounds is greatly affected
by pH. If acid herbicides are to be extracted from TCLP leachates or other samples,
adjust the pH to 1.0 before extraction.
Analvte Group
Extraction pH
Phthalate esters
5-7
Organochlorine pesticides
5-9
Polychlorinated biphenyls (PCBs)
5-9
Organophosphorus pesticides
as received
Nitroaromatics and nitramines
as received
TCLP leachates containing organochlorine pesticides
as produced by TCLP
TCLP leachates containing semivolatiles
as produced by TCLP
TCLP leachates containing phenoxyacid herbicides
1.0
7.3 Setting up the extraction apparatus
7.3.1 Assemble a manifold for multiple disk extractions (Figure 1) using 47-mm
or 90-mm extraction disks. Use a filter flask with the standard filter apparatus for single
extractions. The solid-phase disks that are generally appropriate for each group of analytes
are listed below, and in Table 1.
Analvte Group	Disk Medium
Phthalate esters	C18
Organochlorine pesticides	C18
Polychlorinated biphenyls (PCBs)	C18
Organophosphorus pesticides	SDB-RPS
Nitroaromatics and nitramines	SDB-RPS
TCLP leachates containing organochlorine pesticides	SDB-XC
TCLP leachates containing semivolatiles	SDB-XC
TCLP leachates containing phenoxyacid herbicides	SDB-XC
For nitroaromatics and nitramines, samples also may be extracted using an SPE
cartridge. Assemble the cartridge apparatus according to the manufacturer's instructions,
using Porapak R, or equivalent, SPE cartridges, and proceed to Sec. 7.6.
CD-ROM
3535A - 8
Revision 1
January 1998

-------
7.3.2 If samples contain significant quantities of particulates, the use of a filter aid
or prefilter is advisable for disk extractions. Empore™ Filter Aid 400, Whatman GMF 150,
or equivalent prefilters are recommended.
7.3.2.1	Pour about 40 g of Filter Aid 400 onto the surface of the disk after
assembling the standard filter apparatus.
7.3.2.2	Alternatively, place the Whatman GMF 150 on top of the
extraction disk prior to clamping the glass reservoir into the standard filter apparatus.
7.3.2.3	Do not add the filter aid if using the cartridge extraction procedure
for nitroaromatics and nitramines.
7.4 Washing the extraction apparatus
Prior to use, the extraction disks must undergo two separate washing steps, usually with
different solvents. The steps involved in washing the extraction apparatus before use depend on
the analytes of interest and the sample matrix.
Wash the extraction apparatus and disk with the volume of the solvent listed above by rinsing
the solvent down the sides of the glass reservoir. Pull a small amount of solvent through the
disk with a vacuum. Turn off the vacuum and allow the disk to soak for about one minute.
Pull the remaining solvent through the disk and allow the disk to dry.
7.4.1.1	When using a filtration aid, adjust the volume of all wash solvents
so the entire filtration bed is submerged.
7.4.1.2	In subsequent conditioning steps, volumes should be adjusted
so that a level of solvent is always maintained above the entire filter bed.
7.4.1 First washing step
The following table illustrates the solvents recommended for the first washing step.
Analvte Group
Phthalate esters
Organochlorine pesticides
Polychlorinated biphenyls (PCBs)
Organophosphorus pesticides
Nitroaromatics and nitramines
TCLP leachates containing organochlorine pesticides
TCLP leachates containing semivolatiles
TCLP leachates containing phenoxyacid herbicides
1st solvent wash volume
20 ml_ methylene chloride
20 ml_ methylene chloride
20 ml_ methylene chloride
5 ml_ acetone
5 ml_ acetonitrile
5 ml_ acetone
5 ml_ acetone
5 ml_ acetonitrile
CD-ROM
3535A - 9
Revision 1
January 1998

-------
7.4.2 Second washing step
The following table illustrates the solvents recommended forthe second washing step.
Analvte Group
Phthalate esters
Organochlorine pesticides
Polychlorinated biphenyls (PCBs)
Organophosphorus pesticides
Nitroaromatics and nitramines
TCLP leachates containing organochlorine pesticides
TCLP leachates containing semivolatiles
TCLP leachates containing phenoxyacid herbicides
2nd solvent wash volume
10 ml_ acetone
10 ml_ acetone
not required
5 ml_ methanol
15 ml_ acetonitrile
5 ml_ ethyl acetate
5 ml_ ethyl acetate
not required
7.5 Disk conditioning
The extraction disks are composed of hydrophobic materials which will not allow water to pass
unless they are pre-wetted with a water-miscible solvent before being used for sample extraction.
This step is referred to as conditioning, and the solvent used is dependent on the analytes of
interest. The following table illustrates the solvents recommended for specific groups of analytes.
NOTE: Beginning with the conditioning step, it is CRITICAL that the disk NOT go dry until after
the extraction steps are completed. Should a disk accidentally go dry during the
conditioning steps, the conditioning steps for that disk must be repeated prior to adding
the sample.
Analvte Group	Conditioning steps
Phthalate esters	20 mL methanol, soak 1 min,
20 mL reagent water
Organochlorine pesticides
20 mL methanol, soak 1 min,
20 mL reagent water
Polychlorinated biphenyls (PCBs)
20 mL methanol, soak 1 min,
20 mL reagent water
Organophosphorus pesticides
5 mL methanol, soak 1 min,
20 mL reagent water
Nitroaromatics and nitramines
15 mL acetonitrile, soak 3
min
30 mL reagent water
TCLP leachates containing organochlorine pesticides 5 mL methanol soak 1 min,
15 mL reagent water
CD-ROM
3535A- 10
Revision 1
January 1998

-------
Analvte Group
Conditioning steps
TCLP leachates containing semivolatiles	5 mL methanol soak 1 min,
15 mL reagent water
TCLP leachates containing phenoxyacid herbicides 5 mL methanol soak 1 min,
15 mL reagent water
7.5.1	Add the conditioning solvent to the extraction apparatus. Apply a vacuum
until a few drops of solvent pass through the disk, ensuring that the disk is soaked with the
solvent. Turn off the vacuum and allow the disk to soak in the solvent for the time specified
above.
7.5.2	When using a filtration aid, adjust the volume of conditioning solvents so that
the entire filtration bed remains submerged until the extraction is completed.
7.5.3	Once the soaking time is over, apply the vacuum again, drawing all but a thin
layer of solvent through the disk. Stop the vacuum just before the disk goes dry.
7.5.4	Add the volume of organic-free reagent water listed above and apply vacuum
to draw the water through the disk. Stop the vacuum just before the disk goes dry, leaving
2-3 mm of water above the surface of the disk.
7.5.5	Proceed to Sec. 7.7 for the sample extraction instructions.
7.6 Cartridge procedure for nitroaromatics and nitramines
Aqueous samples to be analyzed for nitroaromatics and nitramines may also be extracted
using the SPE cartridge technique described below. The same sample preparation considerations
discussed in Sec. 7.1 also apply to this procedure.
7.6.1	After assembling the SPE cartridge in the extraction apparatus (see Sec.
7.3.1), wash the cartridge with 10 mL of acetonitrile, using gravity flow. Do not allow the
cartridge to go dry.
7.6.2	When only a thin layer of solvent remains above the sorbent bed in the
cartridge, add 30 mL of reagent water to the cartridge and allow it to flow through the sorbent
bed under gravity flow. Stop the flow just before the cartridge goes dry.
7.6.3	Attach a connector to the top of the cartridge. The other end of the
connector should be fitted with flexible PTFE tubing long enough to reach into the sample
bottle or other container (e.g., a beaker) holding the sample.
7.6.4	Turn on the vacuum, and draw the sample through the cartridge at a rate of
about 10 mL/min, until all of the sample has passed through the cartridge. As particulate
matter plugs the cartridge and slows the flow, increase the vacuum to maintain a reasonable
flow rate.
7.6.5	Once all of the sample has been pulled through the cartridge, shut off the
vacuum and add 5 mL of reagent water to the cartridge. Allow the reagent water to pass
CD-ROM
3535A - 11
Revision 1
January 1998

-------
through the cartridge under gravity flow, if practical, or apply a vacuum to complete the
process. Shut off the flow once the water has been drawn through the cartridge.
7.6.6	Method blanks and matrix spike aliquots (Sec. 7.1) are handled in the same
manner as the samples.
7.6.7	Proceed with sample elution, as described in Sec. 7.9.
7.7	Sample extraction using SPE disks
7.7.1	Pour the sample into the reservoir and, under full vacuum, filter it as quickly
as the vacuum will allow (at least 10 minutes). Transfer as much of the measured volume
of water as possible.
NOTE: With heavily particle-laden samples, allow the sediment in the sample to settle and
decant as much liquid as is practical into the reservoir. After most of the aqueous
portion of the sample has passed through the disk, swirl the portion of the sample
containing sediment and add it to the reservoir. Use additional portions of organic-
free reagent water to transfer any remaining particulates to the reservoir.
Particulates must be transferred to the reservoir before all of the aqueous sample
has passed through the disk.
7.7.2	After the sample has passed through the solid-phase media, dry the disk by
maintaining vacuum for about 3 minutes. Method blanks and matrix spike aliquots (Sec. 7.1)
are handled in the same manner as the samples.
7.8	Elution of the analytes from the disk
The choice of elution solvent is critical to the success of solid-phase extraction. The
recommended elution solvent for each group of analytes is listed below.
Analvte Group
Phthalate esters
Organochlorine pesticides
Polychlorinated biphenyls (PCBs)
Organophosphorus pesticides
Nitroaromatics and nitramines
TCLP leachates containing organochlorine
pesticides
Sample elution steps
5 ml_ acetone, soak 15-20 sec. Rinse bottle
with 15 ml_ acetonitrile and add to disk.
5 ml_ acetone, soak 15-20 sec. Rinse bottle
with 15 ml_ methylene chloride and add to disk.
5 ml_ acetone, soak 15-20 sec. Rinse bottle
with 20 ml_ acetonitrile and add to disk.
0.6 ml_ acetone, soak 1 min. Rinse bottle with
5 ml_ MTBE and add to disk. Repeat bottle
rinse twice more.
5 ml_ acetonitrile, soak 3 min.
Rinse bottle with 4 ml_ acetone and add to
disk. Rinse glassware with 2 ml_ acetone and
add to disk. Soak 1 min. Rinse bottle twice
with 5 ml_ ethyl acetate and add to disk.	
CD-ROM
3535A- 12
Revision 1
January 1998

-------
Analvte Group
TCLP leachates containing semivolatiles
Rinse bottle with 4 mL acetone and add to
disk. Rinse glassware with 2 mL acetone and
add to disk. Soak 1 min. Rinse bottle twice
with 5 mL ethyl acetate and add to disk.
Rinse bottle with 5 mL acetonitrile and add to
disk. Soak 1 min. Rinse bottle twice more with
5 mL acetonitrile and add to disk.
Sample elution steps
TCLP leachates containing phenoxyacid
herbicides
7.8.1	Remove the entire standard filter assembly (do not disassemble) from the
manifold and insert a collection tube. The collection tube should have sufficient capacity to
hold all of the elution solvents. The drip tip of the filtration apparatus should be seated
sufficiently below the neck of the collection tube to prevent analyte loss due to splattering
when vacuum is applied. When using a filter flask for single extractions, empty the water
from the flask before inserting the collection tube.
7.8.2	An initial elution with a water-miscible solvent, i.e., acetone or acetonitrile,
improves the recovery of analytes trapped in water-filled pores of the sorbent. Use of a
water-miscible solvent is particularly critical when methylene chloride is used as the second
elution solvent. With the collection tube in place, add the volume of elution solvent listed
above to the extraction apparatus. Allow the solvent to spread out evenly across the disk
(or inert filter) then quickly turn the vacuum on and off to pull the first drops of sovlent through
the disk. Allow the disk to soak for 15 to 20 seconds before proceeding to Sec. 7.8.3
7.8.3	Rinse the sample bottle and/or glassware that held the sample with the
second solvent listed above and transfer the solvent rinse to the extraction apparatus. As
needed, use a disposable pipette to rinse the sides of the extraction apparatus with solvent
from the bottle.
7.8.4	Draw about half of the solvent through the disk and then release the vacuum.
Allow the remaining elution solvent to soak the disk and particulates for about one minute
before drawing the remaining solvent through the disk under vacuum. When using a filtration
aid, adjust the volume of elution solvent so that the entire filtration bed is initially submerged.
7.8.5	Repeat the bottle rinsing step as listed in the table above, continuing to apply
vacuum and collecting the solvent in the tube.
7.9	Eluting the nitroaromatics and nitramines from the cartridge
Once the reagent water has passed through the column, place a collection tube under the
cartridge. Add 5 mL of acetonitrile to the top of the cartridge and allow it to pass through the
cartridge under gravity flow, collecting the solvent in the collection tube. Measure the volume of
acetonitrile recovered from the cartridge.
7.10	K-D concentration technique
Where necessary to meet the sensitivity requirements, sample extracts may be concentrated
to the final volume necessary for the determinative method and specific application, using the K-D
technique or nitrogen evaporation.
CD-ROM
3535A- 13
Revision 1
January 1998

-------
7.10.1	Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10-mL
concentrator tube to an appropriately sized evaporation flask.
7.10.2	Dry the combined extracts in the collection tube (Sees. 7.8 and 7.9) by
passing them through a drying column containing about 10 g of anhydrous sodium sulfate.
Collect the dried extract in the K-D concentrator. Use acidified sodium sulfate (see Method
8151) if acidic analytes are to be measured.
7.10.3	Rinse the collection tube and drying column into the K-D flask with an
additional 20-mL portion of solvent in order to achieve a quantitative transfer.
7.10.4	Add one or two clean boiling chips to the flask and attach a three-ball Snyder
column. Attach the solvent vapor recovery glassware (condenser and collection device, see
Sec. 4.6) to the Snyder column of the K-D apparatus, following the manufacturer's
instructions. Pre-wet the Snyder column by adding about 1 ml_ of methylene chloride (or
other suitable solvent) to the top of the column. Place the K-D apparatus on a hot water bath
(15 - 20°C above the boiling point of the solvent) so that the concentrator tube is partially
immersed in the hot water and the entire lower rounded surface of the flask is bathed with hot
vapor. Adjust the vertical position of the apparatus and the water temperature as required
to complete the concentration in 10-20 minutes. At the proper rate of distillation the balls
of the column will actively chatter, but the chambers will not flood. When the apparent
volume of liquid reaches 1 ml_, remove the K-D apparatus from the water bath and allow it
to drain and cool for at least 10 minutes.
7.10.4.1	If a solvent exchange is required (as indicated in Table 1),
momentarily remove the Snyder column, add 50 ml_ of the exchange solvent and a
new boiling chip.
7.10.4.2	Reattach the Snydercolumn. Concentrate the extract, raising the
temperature of the water bath, if necessary, to maintain a proper distillation rate.
7.10.5	Remove the Snydercolumn. Rinse the K-D flask and the lower joints of the
Snyder column into the concentrator tube with 1 - 2 ml_ of solvent. The extract may be further
concentrated by using one of the techniques outlined in Sec. 7.11, or adjusted to a final
volume of 5.0 -10.0 ml_ using an appropriate solvent (Table 1).
7.11 If further concentration is required, use either the micro-Snyder column technique
(7.11.1) or nitrogen evaporation technique (7.11.2).
7.11.1 Micro-Snyder column technique
7.11.1.1 Add a fresh clean boiling chip to the concentrator tube and attach
a two-ball micro-Snyder column directly to the concentrator tube. Attach the solvent
vapor recovery glassware (condenser and collection device) to the micro-Snyder
column of the K-D apparatus, following the manufacturer's instructions. Pre-wet the
Snydercolumn by adding 0.5 ml_ of methylene chloride or the exchange solvent to the
top of the column. Place the micro-concentration apparatus in a hot water bath so
that the concentrator tube is partially immersed in the hot water. Adjust the vertical
position of the apparatus and the water temperature, as necessary, to complete the
CD-ROM
3535A- 14
Revision 1
January 1998

-------
concentration in 5 - 10 minutes. At the proper rate of distillation the balls of the
column will actively chatter, but the chambers will not flood.
7.11.1.2 When the apparent volume of liquid reaches 0.5 ml_, remove the
apparatus from the water bath and allow it to drain and cool for at least 10 minutes.
Remove the Snyder column and rinse its lower joints into the concentrator tube with
0.2 ml_ of solvent. Adjust the final extract volume to 1.0 - 2.0 ml_.
7.11.2 Nitrogen evaporation technique
7.11.2.1	Place the concentrator tube in a warm bath (30°C) and evaporate
the solvent volume to 0.5 ml_ using a gentle stream of clean, dry nitrogen (filtered
through a column of activated carbon).
CAUTION: New plastic tubing must not be used between the carbon trap and the
sample, since it may introduce phthalate interferences.
7.11.2.2	Rinse down the internal wall of the concentrator tube several
times with solvent during the concentration. During evaporation, position the
concentrator tube to avoid condensing water into the extract. Under normal
procedures, the extract must not be allowed to become dry.
CAUTION: When the volume of solvent is reduced below 1 ml_, some semivolatile
analytes such as cresols may be lost.
7.12 The extract may now be subjected to cleanup procedures or analyzed for the target
analytes using the appropriate determinative technique(s). If further handling of the extract will not
be performed immediately, stopper the concentrator tube and store in a refrigerator. If the extract
will be stored longer than 2 days, it should be transferred to a vial with a PTFE-lined screw-cap,
and labeled appropriately.
8.0 QUALITY CONTROL
8.1	Any reagent blanks or matrix spike samples should be subjected to exactly the same
analytical procedures as those used for actual samples.
8.2	Refer to Chapter One for general quality control procedures and Method 3500 for
specific QC procedures for extraction and sample preparation.
9.0 METHOD PERFORMANCE
Refer to the determinative methods listed in Sec. 1.1 for performance data related to solid-
phase extraction.
CD-ROM
3535A- 15
Revision 1
January 1998

-------
10.0 REFERENCES
1.	Lopez-Avila, V., Beckert, W., et. al., "Single Laboratory Evaluation of Method 8060 -
Phthalate Esters", EPA/600/4-89/039.
2.	Tomkins, B.A., Merriweather, R., et. al., "Determination of Eight Organochlorine Pesticides
at Low Nanogram/Liter Concentrations in Groundwater Using Filter Disk Extraction and Gas
Chromatography", JAOAC International, 75(6), pp. 1091-1099 (1992).
3.	Markell, C., "3M Data Submission to EPA," letter to B. Lesnik, June 27, 1995.
4.	Jenkins, T. F., Thorne, P. G., Myers, K. F., McCormick, E. F., Parker, D. E., and B. L.
Escalon (1995). Evaluation of Clean Solid Phases for Extraction of Nitroaromatics and
Nitramines from Water. USACE Cold Regions Research and Engineering Laboratory,
Special Report 95-22.
CD-ROM
3535A- 16
Revision 1
January 1998

-------
TABLE 1
SPECIFIC EXTRACTION CONDITIONS FOR VARIOUS DETERMINATIVE METHODS
Determinative Method
Extraction pH
Disk Medium3
Elution Solvent
Exchange Solvent
Final Extract
Volume for
Analysis (ml_)b
8061 (phthalate esters)
8081	(organochlorine
pesticides)
8082	(PCBs)
8141 (organophosphorus
pesticides)
8330 (nitroaromatics and
nitramines)
TCLP pesticides (8081)
TCLP semivolatiles (8270)
TCLP phenoxyacid
herbicides (8321)
5-7
5-9
5-9
as received
as received
as produced by TCLP
as produced by TCLP
1.0
'18
'18
C-I8
SDB-RPS
SDB-RPS
SDB-XC
SDB-XC
SDB-XC
acetonitrile
methylene chloride
methylene chloride
MTBE
acetonitrile
ethyl acetate
ethyl acetate
acetonitrile
hexane
hexane
hexane
hexane
acetonitrile
hexane
methylene chloride
hexane
10.0
10.0
10.0
10.0
10.0
10.0
1.0
10.0
a SDB has a greater capacity than C18 and a greater affinity for more analytes but they may be more difficult to elute.
b For methods where the suggested final extract volume is 10.0 mL, the volume may be reduced to as low as 1.0 mL to achieve lower detection
limits.
CD-ROM
3535A- 17
Revision 1
January 1998

-------
— 250

I —

	20C
*



, j ^150



1 	 100

mm
s
Reservoir
Clamp
.TM
Empore
Extraction Disk
Base
(Fritted or with Screen)
Drip Tube
Filter Flask or Manifold
FIGURE 1
DISK EXTRACTION APPARATUS
CD-ROM
3535A- 18
Revision 1
January 1998

-------
METHOD 3535A
SOLID-PHASE EXTRACTION (SPE)
CD-ROM
3535A- 19
Revision 1
January 1998

-------