US Environmental Protection Agency
Office of Pesticide Programs

Office of Pesticide Programs

Mic ro bio lo gy La bo ra to ry

Environmental Science Center, Ft. Meade, MD

Standard Operating Procedure for

Use and Maintenance of Spectrophotometers

SOP Number: EQ-04-06
Date Revised: 09-02-14


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 1 of 8

SOP Number

EQ-04-06

Title

Use and Maintenance of Spectrophotometers

Scope

This protocol describes the procedure for verifying the performance
of spectrophotometers.

Application

This protocol describes the procedure for determining the absorbance
and percent transmittance of microbial suspensions at a wavelength
of 650 nm.





Approval Date

SOP Developer:



Print Name:

SOP Reviewer



Print Name:

Quality Assurance Unit



Print Name:

Branch Chief



Print Name:





Date SOP issued:



Controlled copy
number:



Date SOP withdrawn:




-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 2 of 8

TABLE OF CONTENTS
Contents	Page Number

1.

DEFINITIONS

3

2.

HEALTH AND SAFETY

3

3.

PERSONNEL QUALIFICATIONS AND TRAINING

3

4.

INSTRUMENT CALIBRATION

3

5.

SAMPLE HANDLING AND STORAGE

4

6.

QUALITY CONTROL

5

7.

INTERFERENCES

5

8. NON-CONFORMING DATA

5

9.

DATA MANAGEMENT

5

10.

CAUTIONS

5

11.

SPECIAL APPARATUS AND MATERIALS

6

12.

PROCEDURE AND ANALYSIS

6

13.

DATA ANALYSIS/CALCULATIONS

8

14.

FORMS AND DATA SHEETS

8

15.

REFERENCES

8


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 3 of 8

1. Definitions

Additional abbreviations/definitions are provided in the text.

1.	A = Absorbance, also referred to as optical density (OD). The measure of
the quantity of light that a sample neither transmits nor reflects and is
proportional to the concentration of a substance in a solution. The
absorbance scale is logarithmic. In turbid bacterial cultures, the scatter of
light is measured; this scatter is what the spectrophotometer interprets as
absorbance.

2.	%T = Percent transmittance. The ratio of the intensity of the light that has
passed through the sample to the intensity of the light when it entered the
sample (T = lout / Iin). The transmittance is displayed as a percentage on an
analog scale.

NOTE: To convert between the absorbance and transmittance scales, use
the equation: A = -log (%T/100). Values for both A and %T are unitless
as both are derived from a ratio between the incident light and the
transmitted light.

2. Health and
Safety

1.	Follow procedures specified in SOP MB-01, Laboratory Biosafety. The
Study Director and/or lead analyst should consult the Material Safety Data
Sheet for specific hazards associated with products.

2.	Perform all M bovis (BCG) manipulations in room B202 or room B207.

3.	Cap cuvettes containing microbial suspensions prior to removal from the
BSC. The use of a cuvette rack for transporting capped cuvettes within
the laboratory is recommended. For transporting microbial suspensions in
capped cuvettes between laboratories, place the cultures in secondary
containment.

3. Personnel
Qualifications
and Training

Refer to SOP ADM-04, OPP Microbiology Laboratory Training.

4. Instrument
Calibration

1.	Vendor calibration/certification.

a. Use an ISO 17025 accredited vendor to certify the
spectrophotometers on an as-needed basis.

2.	In-house verification of spectrophotometers.

a.	Check both spectrophotometers on a quarterly basis using
spectrophotometer standards (e.g., In-Spec Spectrophotometer
standards and background solution) aliquoted into optical glass
cuvettes.

b.	Use three spectrophotometer reference standards (a 0.3, 0.5, and 0.8
absorbance standard) to verify the performance of the Denville®


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 4 of 8



S800 Spectrawave.

c.	Use four spectrophotometer reference standards (a 0.3, 0.5, and 0.8
absorbance standard and a 20% transmittance standard) to verify the
performance of the Beckman DU 730.

d.	Warm each instrument for at least 1 hour prior to verification (refer
to 15.3).

e.	Do not shake, mix, or agitate the standards. Discard any used or
unused aliquoted standards.

f.	Zero the instrument using the background solution to blank the
system at 650 nm.

g.	Wipe both sides of the cuvette before and after putting the
background solution in the cuvette.

h.	Exchange the contents in the sample cuvette with one reference
standard solution. Be sure to rinse the cuvette several times with the
standard.

i.	Begin with the least concentrated standard solution. Position the
cuvette using the same orientation for each measurement. After
reading the least concentrated standard solution, exchange the
contents of the cuvette with the next standard following the same
procedure as in 4.2f.

j. Compare the data generated by the instrument with the certified
values provided with each standard. The acceptable tolerance for
each instrument is as follows:

Tolerance = Uncertainty + Instrument's Photometric Accuracy

k. The photometric accuracy for the Beckman DU 730 is ±0.005 A at
0.0 to 0.5 A, 1% at 0.5 to 2.0 A.

1. The photometric accuracy for the Denville® S800 Spectrawave is
±0.003 A at 0.0 to 0.5 A (at 546 nm).

m. Clean optical glass cuvettes using alcohol and de-ionized water after
use.

5. Sample

Handling and
Storage

1.	For sample readings, aliquot microbial suspensions into cuvettes in the
biological safety cabinet (BSC) and cap prior to removal from the BSC.
Cuvettes should remain capped at all times once removed from the BSC.

2.	Never leave cuvettes containing microbial suspensions in the cell holder
once a reading has been taken.


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 5 of 8



3.	Fill all blank and sample cuvettes at least half-full to a level just above the
window shoulder (23 mm). Once filled, cap the cuvette. Avoid touching
the cuvettes below the shoulder.

4.	Check that no air bubbles are present in the cuvette; gently tap to remove
if necessary. Using a clean, dry, and lint free towel, gently wipe the
cuvette sides oriented in the optical path (indicated with an arrow - see

11.3) and firmly position the cuvette in the sample holder.

6. Quality Control

1.	For quality control purposes, the required information is documented on
the appropriate form(s) (see section 14).

2.	Certification records are maintained in D217.

7. Interferences

1. The entire sampling module must be kept clean. If it becomes dirty, wipe
it clean with soap and water and a soft cloth.

8. Non-
conforming
Data

1.	Management of non-conforming data will be consistent with SOP ADM-
07, Non-Conformance Reports.

2.	For the Beckman DU 730, if any function does not pass or if any error
messages are displayed, refer to section 17, Troubleshooting in the
Operating Instructions (see 15.1) and, if necessary, contact the
manufacturer for assistance. Record corrective actions on the Systems
Check Record Form under "Comments." Remove the spectrophotometer
from service if troubleshooting does not resolve the issue.

3.	For the Denville® S800 Spectrawave, refer to the Error Messages section
of the user manual to troubleshoot if FAIL appears on the instrument (see
15.2).

9. Data

Management

Archive Data consistent with SOP ADM-03, Records and Archives. The
calibration certificates are filed and maintained in a secured file cabinet in the
file room D217. Only authorized personnel have access to the secured files.
Archived data is subject to OPP's official retention schedule contained in SOP
ADM-03, Records and Archives.

10. Cautions

1.	Maintain an ambient temperature of 10-35°C for proper instrument
operation. Leave at least a 15 cm clearance at the top and on all sides for
air circulation. Keep the air vents under the instrument clean and free of
materials that might obstruct the air flow.

2.	Allow the instrument to warm up for a minimum of 5 minutes before
taking any sample readings (for instrument verification, allow the
instrument to warm up for at least 1 hour).

3.	During system startup and initialization for the Beckman DU 730, the
analyst must remain present during the entire system check of the


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 6 of 8



instrument and must observe the check marks in order to verify that the
instrument has passed its system checks.

4.	During system startup and initialization of the Denville® S800
Spectrawave, the unit will briefly display a "PASS" statement on the
machine. If the Denville® S800 Spectrawave does not pass, the word
"FAIL" will appear on the display together with a symbol and a number.

5.	Do not switch the instrument off and on in rapid succession; always wait
at least 5 seconds before switching the instrument on again.

6.	Lenses are located on both sides of the sampling module. Wipe lenses
with a soft lint-free and oil free cloth to clean. Do not use organic solvents
such as acetone to clean the lenses.

7.	Due to its shorter optical path, do not use the Denville® S800 Spectrawave
to measure the optical density of Mycobacterium bovis (BCG) or
Mycobacterium terrae.

11. Special

Apparatus and
Materials

1.	Beckman DU Series 730 Life Science Spectrophotometer: wavelength
range 190-1100 nm.

a. The Beckman DU 730 operates in both the visible and UV range.
Use cuvettes made of silica or quartz in the UV region (190-360 nm).
Use glass or plastic cuvettes in the visible region (e.g., 650 nm). To
accurately measure turbid samples like bacterial cultures, a turbidity
cell holder (part number A23623) is installed.

2.	The Denville® S800 Spectrawave operates in the visible region (330-800
nm). The optical height is 15 mm; the minimum volume that can be used
is 0.7 mL in a semi-micro cuvette.

3.	Cuvettes.

a.	Semi-micro plastic disposable cuvettes: 12.5 x 12.5 x 45 mm,
capacity of 1.5-3 mL, 10 mm light path. A triangular arrow at the top
of each cuvette indicates the optical path orientation.

b.	Optical glass cuvettes. 10 mm light path; used for spectrophotometer
standards.

12. Procedure and
Analysis



12.1 BeckmanDU
730: Self-
Check

a.	Close the cell compartment and turn on the instrument.

b.	Each time the instrument is powered up, a series of diagnostic tests is
performed automatically to ensure operation of major system
components. This procedure, which takes two minutes, checks the
system, lamps, wavelength calibration, filter adjustment and voltage.


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 7 of 8



The instrument also contains programs for checking photometric
accuracy, photometric noise, stray light, wavelength accuracy, lamp
history, and printer check.

c.	After each portion of the self-check, a check mark will appear next to
each diagnostic test.

NOTE: Analyst must be present during the instrument's self-check
(see section 10.3).

d.	When the power-up diagnostics are complete, the Main Menu
appears. If any function does not pass or if any error messages
appear during the self-check, refer to the "Troubleshooting" section
of the manual (see ref. 15.1).

e.	Record results of the system's self-check on the appropriate form
(see section 14).

12.2 Using the
Beckman DU
730

a.	From the Main Menu, select User Programs on the touch screen,
select the desired program (e.g., Fixed Wavelength) and touch Start
to run the program.

b.	Touch X to enter the desired wavelength (e.g., 650 nm).

c.	Before taking measurements, blank the instrument using a cuvette
with a "blank" solution, such as the growth medium or other
appropriate background liquid.

d.	Insert the blank cuvette into the cell holder and close the cell
compartment. Touch Blank to take the blank reading. When done,
the instrument enables the Read button.

e.	Remove the blank and insert a sample cuvette into the cell holder and
close the cell compartment. Touch Read to take the sample reading.

f.	To toggle between absorbance and percent transmittance readings,
touch the Abs/%Trans button.

g.	For multiple sample readings, it is advisable to take a blank reading
between samples.

h.	Refer to the manual (see ref. 15.1) when using other programs.

12.3 Denville®
S800

Spectrawave -
Self-Check

a.	Each time the instrument is powered up, a series of diagnostic tests is
performed automatically to ensure operation of major system
components including lamp performance and wavelength calibration.

b.	After the self-check, a "PASS" will briefly appear on the
instrument's display. If the instrument does not pass or if any error


-------
SOP No. EQ-04-06
Date Revised 09-02-14
Page 8 of 8



messages appear during the self-check, refer to the
"Troubleshooting" section of the manual (see ref. 15.2).

c. Record results of the system's self-check on the appropriate form
(see section 14).

12.4 Using the
Denville®
S800

Spectrawave

a.	Using the arrow keys, highlight nm on the display and press the red
•S button. Select the desired wavelength using the arrow keys then
press the red S button.

b.	Before taking measurements, blank the instrument using a cuvette
with a "blank" solution, such as the growth medium or other
appropriate background liquid.

c.	Insert the blank cuvette into the cell. Press the blue R button to take
the blank reading.

d.	Remove the blank and insert a sample cuvette into the cell holder.
Press the green T button to take the sample reading.

e.	To toggle between absorbance and percent transmittance readings,
use the arrows to highlight Abs/%T on the display and then use the
¦w A keys to select absorbance or percent transmittance.

13. Data Analysis/
Calculations

1. None

14. Forms and Data
Sheets

Test Sheets. Test sheets are stored separately from the SOP under the
following file names:

Spectrophotometer Systems Check Record EQ-04-06 F1 .docx

Spectrophotometer Calibration Verification EQ-04-06_F2.docx

15. References

1.	Operating Instructions: Beckman DU Series 700 Spectrophotometer
User's Guide. Beckman Instructions A24014-AB, July 2007.

2.	Denville® S800 Spectawave User Manual. Issue 02 - 03/2005.

3.	Scanning Procedures for Photometric Standards. (2012). Retrieved June
10, 2014, from http://www.in-spec.com/index.php?id=2170.


-------