US Environmental Protection Agency Office of Pesticide Programs Office of Pesticide Programs Microbiology Laboratory Environmental Science Center, Ft. Meade, MD Standard Operating Procedure for Calibration and Maintenance of Weigh Balances SOP Number: EQ-03-07 Date Revised: 01-29-14 ------- SOP No. EQ-03-07 Date Revised 01-29-14 Page 1 of 6 SOP Number EQ-03-07 Title Calibration and Maintenance of Weigh Balances Scope Describes process for use, calibration, and quality control of weigh balances and reference weights. Application Weigh balances are used to measure the weight of objects in the laboratory, such as media and reagent ingredients and disinfectant containers. Reference weights are used as reference standards to verify the calibration of the weigh balances. Approval Date SOP Developer: Print Name: SOP Reviewer Print Name: Quality Assurance Unit Print Name: Branch Chief Print Name: Date SOP issued: Controlled copy number: Date SOP withdrawn: ------- SOP No. EQ-03-07 Date Revised 01-29-14 Page 2 of 6 TABLE OF CONTENTS Contents Page Number 1. DEFINITIONS 3 2. HEALTH AND SAFETY 3 3. PERSONNEL QUALIFICATIONS AND TRAINING 3 4. INSTRUMENT CALIBRATION 3 5. SAMPLE HANDLING AND STORAGE 3 6. QUALITY CONTROL 3 7. INTERFERENCES 3 8. NON-CONFORMING DATA 3 9. DATA MANAGEMENT 3 10. CAUTIONS 4 11. SPECIAL APPARATUS AND MATERIALS 4 12. PROCEDURE AND ANALYSIS 4 13. DATA ANALYSIS/CALCULATIONS 6 14. FORMS AND DATA SHEETS 6 15. REFERENCES 6 ------- SOP No. EQ-03-07 Date Revised 01-29-14 Page 3 of 6 1. Definitions 1. ISO = International Organization for Standardization 2. Tolerance for weigh balances = Acceptable limits or range in measurement (in grams) that the laboratory can tolerate. See 12.4. 2. Health and Safety None 3. Personnel Qualifications and Training Refer to SOP ADM-04, OPP Microbiology Laboratory Training. 4. Instrument Calibration 1. Weigh balances are inspected, cleaned and calibrated annually by an ISO 17025 accredited vendor. 2. Calibration of the reference weights is performed every two years by an ISO 17025 accredited vendor. Weights may be calibrated more frequently if deemed necessary (e.g., weight is dropped, chipped, etc.). 5. Sample Handling and Storage 1. Wear clean cotton gloves (supplied with reference weights) or use forceps while handling reference weights. To avoid depositing oil and dirt onto the surface of the weight, do not touch weights with bare hands. 2. Store reference weights in cases provided by the manufacturer. 6. Quality Control For quality control purposes, the required information is documented on the appropriate form(s) (see section 14). 7. Interferences 1. For optimal performance, place balance on a stable, even, horizontal surface with minimal vibration. Avoid areas with excessive heat and moisture, direct sunlight, aggressive chemical vapors, and drafts. 2. If a balance is transferred to a different location, the accuracy check (section 12.3) must be performed prior to use in the new location. 8. Non- conforming Data 1. When verifying the calibration of weigh balances (section 12.3), any discrepancies in weight measurements will be confirmed by repeating the operation. A service representative will be notified, if necessary, to re- calibrate the instrument when the calibration check shows that the weigh balance is outside of the acceptable tolerance range (see section 12.4). 2. If the vendor determines that a weigh balance or reference weight is out of tolerance and cannot be properly calibrated, the equipment will be replaced. 3. Procedures will be consistent with SOP ADM-07, Non-Conformance Reports. 9. Data Management 1. An inventory of weigh balances and reference weights requiring vendor calibration will be maintained electronically (see section 14). After each ------- SOP No. EQ-03-07 Date Revised 01-29-14 Page 4 of 6 addition to or deletion from the inventory, file a hard copy of the inventory in the Weigh Balance Calibration Record book. 2. Data will be archived consistent with SOP ADM-03, Records and Archives. 10. Cautions 1. Reference weights are removed from service when the calibration expires (two years from the date of calibration) and are returned to service when recalibration is completed. 2. Annual calibration of weigh balances is performed at approximately the same time each year. 3. See section 5 for guidance on proper handling of reference weights. 11. Special Apparatus and Materials 1. Weigh balances. Used to measure the weight of objects in the laboratory, such as media and reagent ingredients and disinfectant containers. 2. Reference weight set (range of lg to 50g) and Individual reference weights (1 mg, 10 mg, 100 mg, 100 g, 500 g, 1 kg, 2 kg). Used as reference standards to verify the calibration of the weigh balances. 12. Procedure and Analysis Calibration certificates must contain the stamp of the accrediting body (e.g., A2LA, NVLAP) and the calibration vendor certificate number. 12.1 Calibration of Weights a. Before the calibration of a weight or weigh set expires (two years from the date of calibration), remove it from service. b. Consult ISO 17025 accredited vendor regarding quote for service, packing/shipping instructions, and completion of any required forms prior to shipping weights. c. Pack and ship weights to vendor. d. Once the weight or weight set has been recalibrated and shipped back to the laboratory, file the calibration certificate in the Weigh Balance Calibration Record book and return the weights to service. 12.2 Daily Calibration and Use of Weigh Balances a. Follow the instructions provided by the manufacturer for the operation of each weigh balance. See section 15. b. Each balance has a built-in calibration system. c. When weighing, apply load to center of balance. Close balance doors, if applicable, to reduce draft. d. Clean balance pan after each use with a soft brush or damp towel. Allow the balance to dry before the next use. 12.3 Six Month Accuracy a. Check each weigh balance, approximately six months following vendor calibration, for accuracy using reference weights. ------- SOP No. EQ-03-07 Date Revised 01 -29-1 4 Page 5 of 6 Check of Weigh Balances b. See the Reference Weight Selection for Sixih Month Aceuiaey Check (section 14) for guidance on which reference weights to use to verify the calibration of each weigh balance. c. See section 5 for guidance on handling reference weights. d. Tare or "zero" the weigh balance before the addition of each weight. e. Add weights to center of balance and close balance doors, if applicable. £ Record results for each reference weight on the Verification of Weigh Balance Calibration Record Form (section 14). g- See section 12.4 for acceptable tolerances for weigh balances when conducting the accuracy check. h Note that the acceptable tolerance varies for the 10 and 100 nig reference weights, depending upon whether they are being weighed on a top loading balance that reads to 0.0lg or a more sensitive analytical balance. 12.4 Six Month Accuracy Cheek of Weigh Balances - Total Load Applied to Weigh Balance Acceptable Tolerance Acceptable Range of Weigh Balance Readings 1 nig ± 0.00005 g 0.00095 g to 0.00105 g Acceptable Tolerances 10 mgA ± 0.0005 g 0.0095 g to 0.0105 g 10 mgB None^ 0.01 g ~~ [00 ulgA ± 0.005 g 0.095 g to 0.105 g 100 mgB ±0.01 g 0.09 g to 0.11 g 1 g ± 0.01 g 0.99 g to 1.01 g 2 g ±0.01 g 1.99 g to 2.01 g 5 g ±0.01 g 4.99 g to 5.01 g 10 g ±0.01 g 9.99 g to 10.01 g 20 g ± 0.01 g 19.99 g to 20.01 g 50 g ± 0.01 g 49.99 g to 50.01 g 100 g ±0.1 g 99.9 g to 100.1 g 150 g ±0.1 g 149.9 g to 150,1 g 500 g ± 0.1 g 499.9 g to 500.1 g ------- SOP No. EQ-03-07 Date Revised 01-29-14 Page 6 of 6 1kg ±1.0g 999 g to 1001 g 2 kg ±1.0g 1999 g to 2001 g 3 kg ±1.0g 2999 g to 3001 g 4 kg ±1.0g 3999 g to 4001 g 5 kg ±1.0g 4999 g to 5001 g AWhen weighed on an analytical balance. BWhen weighed on a top loading balance that reads to O.Olg. 12.5 Annual Calibration of Weigh Balances a. Contact ISO 17025 accredited vendor and schedule a date for calibration. b. Weigh balances are not shipped out. The vendor inspects, cleans, and calibrates balances on site. c. File the calibration certificate in the Weigh Balance Calibration Record book. 13. Data Analysis/ Calculations None 14. Forms and Data Forms are stored separately from the SOP under the following file names: Sheets Sample Inventory of Weigh Balances and EQ-03-07_Fl.docx Reference Weights Requiring Vendor Calibration Reference Weight Selection for Six Month Accuracy Check EQ-03-07_F2.docs Verification of Weigh Balance Calibration Record Form EQ-03-07_F3.docx 15. References Operation manuals for weigh balances located in file cabinet in D-wing. ------- |