US Environmental Protection Agency Office of Pesticide Programs Office of Pesticide Programs Microbiology Laboratory Environmental Science Center, Ft. Meade, MD Standard Operating Procedure for Verification of Digital Pipettes SOP Number: QC-19-06 Date Revised: 01-23-14 ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 1 of 13 SOP Number QC-19-06 Title Verification of Digital Pipettes Scope Describes process for verification of digital pipettes and the calibration of the Artel PCS 2 Pipette Calibration System. Application The Artel PCS 2 pipette calibration system is used to verify volumes of one milliliter or less. The gravimetric procedure is used to verify volumes greater than one milliliter. Approval Date SOP Developer: Print Name: SOP Reviewer Print Name: Quality Assurance Unit Print Name: Branch Chief Print Name: Date SOP issued: Controlled copy number: Date SOP withdrawn: ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 2 of 13 TABLE OF CONTENTS Contents Page Number 1. DEFINITIONS 3 2. HEALTH AND SAFETY 3 3. PERSONNEL QUALIFICATIONS AND TRAINING 3 4. INSTRUMENT CALIBRATION 3 5. SAMPLE HANDLING AND STORAGE 3 6. QUALITY CONTROL 4 7. INTERFERENCES 4 8. NON-CONFORMING DATA 4 9. DATA MANAGEMENT 4 10. CAUTIONS 4 11. SPECIAL APPARATUS AND MATERIALS 5 12. PROCEDURE AND ANALYSIS 5 13. DATA ANALYSIS/CALCULATIONS 9 14. FORMS AND DATA SHEETS 9 15. REFERENCES 9 ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 3 of 13 1. Definitions Abbreviations/definitions are provided in the text. 1. Verification failure = Verification of pipette exceeds ±5% of the target volume. 2. Ideal volume = Target volume; actual volume being measured, corrected for temperature using the density of water at 21.0°C (0.997995 g/mL, see section 15.1). 2. Health and Safety 1. Follow procedures specified in SOP MB-01, Laboratory Biosafety. The Study Director and/or lead analyst should consult the Material Safety Data Sheet for specific hazards associated with products. 2. All kit reagents are non-hazardous, aqueous based and may be disposed of by flushing down a drain. 3. Personnel Qualifications and Training Refer to SOP ADM-04, OPP Microbiology Laboratory Training. 4. Instrument Calibration 1. Refer to SOP EQ-03 (weigh balances) for details on method and frequency of calibration. 2. Check the PCS instrument calibration twice a year using an instrument calibrator kit. 3. Photocopy the PCS instrument calibration printout (it is printed on thermal paper which fades over time) and store in the Pipette Verification and Calibration Log Book. 4. Store certificates of calibration for PCS calibrator kit and PCS Reagent Kit in the Pipette Verification and Calibration Log Book. 5. Acceptable limits for PCS Instrument Calibration are listed in the PCS 2 Pipette Calibration System Procedure Guide (refer to reference 15.2). If a failure occurs, clean the calibration vials and perform the instrument calibration check again. If the instrument fails again, purchase another calibration set and recheck the calibration. If failures persist, schedule service for the instrument and do not use the instrument for calibration until it has been recertified. 6. Refer to Chapter 5 of the PCS Procedure Guide for detailed instructions and pass/fail specifications. 5. Sample Handling and Storage 1. Store the calibration kit and the reagent kit in darkness at room temperature (e.g., 20-25°C) in the same laboratory as the instrument. 2. Handle each blank reagent vial with care. The vial contains a predetermined amount of liquid and any spillage will lead to inaccurate ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 4 of 13 results. 6. Quality Control 1. For quality control purposes, the required information is documented on the appropriate form(s) (see section 14). 2. To provide maximum life and dependable system performance, periodic cleaning should be performed. Refer to Chapter 7, Section 7.3, of the PCS Procedure Guide for instructions on cleaning the cabinet, the vial holder, and the filer wheel assembly. 7. Interferences 1. To ensure that accurate and precise results are obtained during pipette verifications and instrument calibration checks, all vials should be free of fingerprints or smudges prior to placement into the vial block. THUS, AVOID HANDLING THE LOWER PORTION OF THE VIALS. Fingerprints or smudges on the vials that fall into the optical path will introduce errors that cause inaccurate readings. If the lower portion of the vial is inadvertently touched, it should be properly cleaned prior to use (refer to Chapter 2, Section 2.2 of the PCS Procedure Guide for cleaning procedures). 2. Never remove the cap of a calibration standard. 3. Blank reagent vials should be moved immediately from the vial block into the vial holder. Once the blank reagent vial has been inserted into the instrument it must not be turned, moved, or removed until the entire procedure has been completed, or the vial has been consumed. 4. Always take a fresh aliquot of sample solution and lightly cap. Always use the aliquot of sample solution within 30 min to avoid evaporation. 8. Non- conforming Data 1. Management of non-conforming data will be specified in the study protocol; procedures will be consistent with SOP ADM-07, Non- Conformance Reports. 2. Do not use pipettes if the inaccuracy exceeds ±5% of the target volume. 3. See Chapter 7, Troubleshooting, of the PCS Procedure Guide for specific system performance checks. Section 7.2 provides a troubleshooting guide for symptoms, causes, and corrective actions. 9. Data Management 1. Data will be archived consistent with SOP ADM-03, Records and Archives. 2. Maintain an inventory of pipettes electronically using a Microsoft Excel spreadsheet (refer to section 14). After each addition to or deletion from the inventory, file a hard copy of the pipette inventory in the Pipette Verification and Calibration Record Book. 10. Cautions 1. Keep the PCS Instrument out of direct sunlight and away from anything ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 5 of 13 hot or cold. 2. To ensure proper sample mixing, place the PCS Instrument on a sturdy table. 3. Do not open the PCS Vial Holder Cover until you are prompted with "Pipette Sample #" or "Remove Cap and Insert Blank." 4. If a pipette is dropped or damaged, it must be successfully verified or recalibrated prior to use. 5. If a pipette fails the in-house verification assessment, it will not be used in the laboratory and will be recalibrated prior to use. 11. Special Apparatus and Materials 1. ARTEL PCS 2 Pipette Calibration System Instrument, serial no. 6250. For verifying pipettes with volumes 1 mL or less. 2. Calibrated balance. Capable of measuring 0.01 g. For verifying pipettes with volumes greater than 1 mL. 3. Pipettes. a. Rainin Adjustable Volume Pipettes b. Gilson Microman Positive Displacement Pipettes c. Gilson Distriman Continuously Adjustable Volume Repetitive Pipette 12. Procedure and Analysis 12.1 Pipette Verification Requirements a. Twice a year, verify the performance of each pipette listed in Attachment A, once using an in-house procedure (see 12.1, i-ii) and once by a vendor who will also calibrate and service each pipette. i. For pipettes dispensing volumes of 1 mL or less, use the Artel PCS 2 pipette calibration system (refer to section 15.3). ii. For pipettes dispensing volumes greater than 1 mL, use the gravimetric verification procedure. b. If a pipette is out of compliance prior to the biannual verification (e.g., the pipette was dropped), verify the pipette using the Artel PCS 2 pipette calibration system or the gravimetric verification procedure. c. Record the results on the Pipette Verification Record Sheet (refer to section 14). The Pipette Verification Record Sheet is based on the inventory of pipettes and may change over time. Note: after the annual vendor calibration, complete the Verification Status and Corrective Action Taken columns on the Pipette Verification Record ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 6 of 13 Sheet for the half of the year in which the pipettes are calibrated by a vendor. Fill in the Percent Inaccuracy column with "N/A". 12.2 Using the Artel PCS a. Turn on the PCS instrument and the printer. b. Load blank vials into the vial block at least 15 minutes prior to use. Four vials can be loaded into the vial block at this time. Inspect for smudges and clean them if necessary (refer to section 7.1). c. Allow the instrument to self calibrate and follow the on-screen instructions. Enter a pre-assigned Operator ID number. d. When prompted, enter the four or five digit lot code shown on the reagent kit insert and on the exterior label affixed to the top of the reagent kit. If using a new lot, enter the lot information when prompted. e. After self calibration, the instrument will prompt the operator to insert the CAL A Vial. Insert the CAL A calibration standard into the vial holder with the CAL A label facing toward the front of the PCS. f. After closing the cover on the vial holder, the PCS will perform a zero check. Once the zero check is complete, remove the CAL A vial and place it back in the Reagent Kit. g. Remove a blank vial from the block. Invert it several times to thoroughly mix the contents. h. Insert the blank reagent vial into the vial holder located under the vial holder cover. Close the cover. The instrument will take a reading of the blank. i. Enter the pipette serial number (numerical digits only) and the volume to be checked. Once entered, the instrument will indicate which range solution should be used for that verification. i. Refer to "Verification Volume" column in Sample Pipette Verification Sheet for the pipette volumes to be verified. j. Mix the range solution by inverting the bottle several times before removing the cap. Recap the bottle as soon as possible to avoid evaporation. i. Each kit contains bottles of sample solution corresponding to four different pipette volume ranges: Range 1: For pipette volumes 200 to 5000 |iL. Range 2: For pipette volumes 50 to 199 |iL. ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 7 of 13 Range 3: For pipette volumes 10 to 49 |iL. Range 4: For pipette volumes 2 to 9 |iL. ii. Use the proper range solution for the pipette volume being verified. k. Transfer an aliquot of sample solution, using a disposable transfer pipette, into a sample aliquot container. Transfer pipettes and sample aliquot containers are found in the reagent kit. Place the container into the aliquot container holder to avoid spillage. Place the cap lightly on the container to minimize evaporation. 1. Using the pipette to be verified, aspirate an aliquot of sample solution from the sample aliquot container. i. Use standard laboratory pipetting technique and follow specific pipette manufacturer's instructions. m. Lift the instrument vial holder cover and dispense the aliquot of sample solution into the blank reagent vial directly into the liquid or against the inside wall of the vial. If dispensing against the inside wall, place the pipette tip above the meniscus but below the top of the vial holder. The mixing action of the PCS will wash the solution off the inside wall of the vial. n. Once the aliquot is dispensed, carefully withdraw the pipette and close the cover. The PCS will display the results of the previous reading and will prompt for the addition of the next sample. Follow the same above procedure for sample #2 and for each subsequent sample addition. After each new sample addition, the instrument will mix, read, and print the delivered volume. o. After dispensing the desired number of samples for the verification run (5 samples), press the "End of Run" button. The group statistics will be computed and the results printed. p. Within a blank vial's allowed capacity, additional verification runs at any volume may be performed. q. When verifying an adjustable volume pipette at multiple volumes, start anywhere along its volume range. After dispensing the desired number of samples at the first volume, press "Enter" (do not press "End of Run"). The group statistics for the first volume will be computed and the results printed. Follow the on screen instructions and indicate the next volume to be verified when prompted. If desired, continue to verify the same pipette at different volumes by pressing the "Enter" key at the conclusion of each volume. After ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 8 of 13 pipetting the final sample at the last volume for the pipette press the "End of Run" button. The group statistics and results for the last volume and the footer will be printed. 12.3 Gravimetric Verification Procedure a. Record all pertinent information for the gravimetric verification procedure on the Pipette Verification - Gravimetric Analysis Form (refer to section 14). b. In advance of testing, fill a container with de-ionized water and allow it to equilibrate to room temperature in the same laboratory with the balance. List the balance that will be used on the appropriate form. c. Place a small Erlenmeyer flask on the balance and record its mass on the appropriate form. d. Using the pipette to be verified, aspirate an aliquot of DI water from the sample aliquot container and dispense into the Erlenmeyer flask. e. Record the mass on the appropriate form. f. Follow the procedure in 12.3d-e for each subsequent sample addition; measure at least 5 samples. Do not tare between samples. After each new sample addition, record the mass on the appropriate form. g. Input the measurements and other appropriate information into the Pipette Verification - Gravimetric Analysis Spreadsheet. 12.4 Pipette Verification Record Sheet a. From the data on the PCS printout for each pipette or the Pipette Verification - Gravimetric Analysis Spreadsheet, record the following on the Pipette Verification Record Sheet: i. Percent Inaccuracy, record the value from the PCS printout or the Pipette Verification - Gravimetric Analysis Spreadsheet for each pipette. (x — ideal volume) x 100 Percent Inaccuracy = -— ideal volume ii. Verification Status: record as "pass" or "fail". Record status as "pass" if the percent inaccuracy is within ±5%, otherwise record as "fail". iii. Corrective Action: record as "yes" or "no". Record as "yes" if the verification status is "fail" and describe the action taken at the bottom of the page. If the verification status is "pass," record as "no." b. Generate one Pipette Verification Record Sheet biannually and file in ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 9 of 13 the Pipette Verification and Calibration Log Book. c. The Pipette Verification Record Sheet may be completed electronically or by hand. d. Photocopy the PCS printout for each pipette and store in the Pipette Verification and Calibration Log Book. 13. Data Analysis/ Calculations 1. None 14. Forms and Data Sheets 1. Sample Pipette Verification Record Sheet 2. Sample PCS Printout 3. Sample Pipette Inventory 4. Test Sheets. Test sheets are stored separately from the SOP under the following file names: MLB Pipette Inventory QC-19-06 Fl.xlsx Pipette Verification Record Sheet QC-19-06_F2.xlsx Pipette Verification - Gravimetric Analysis Form QC-19-06 F3.xlsx Pipette Verification - Gravimetric Analysis , „ „ , . , c , , A QC-19-06 F4.xlsx Spreadsheet - 15. References 1. CRC Handbook of Chemistry and Physics. 93rd ed. CRC Press: Boca Raton, FL, 2012; p 6-8. 2. Artel, Inc. 1997. PCS 2 Pipette Calibration System Procedure Guide, Doc #: 15A2135 Ver: 5.1, 03/28/97. 3. Artel, Inc. 2003. Artel Lab Report - Calibration Frequency for Pipettes, Issue 6. ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 10 of 13 Sample Pipette Verification Record Sheet OPP Microbiology Laboratory Verification Date(s): Initials: Manufacturer Model No. Serial No(s). Volume Range Verification Volume(s) Percent Inaccuracy1 Verification Status2 Corrective Action (Y/N) X12523D Gilson M10 X12649D 1-10 nL 10 nL BH15232 Gilson M100 GG05125 10-100 nL 100 nL GG05127 1-12.5 nL 10 nL AE10020 10-125 nL 100 nL 100-1250 nL 900 nL 1-12.5 nL 10 nL BD10010 10-125 nL 100 nL 100-1250 nL 900 nL 1-12.5 nL 10 nL U10048H 10-125 nL 100 nL 100-1250 nL 900 nL A0504243A L0508039A C0823986A J0753884A C0825980A Rainin L-1000 C0823596A 100-1000 nL 1000 nL H0101474A H0100977A J0902334A J0908624A D0303509A E0301364A A0510192A Rainin L-200 L0509218A 20-200 nL 100 nL C0821962A J0750805A ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 11 of 13 Manufacturer Model No. Serial No(s). Volume Range Verification Volume(s) Percent Inaccuracy1 Verification Status2 Corrective Action (Y/N) C0820542A C0820661A G0101809A G0102379A J0902612A C0401654A Rainin L-100 C0825210A 10-100 nL 100 nL C0825238A A0507382A C0822081A J0724932A C0822312A Rainin L-20 C0825315A 2-20 nL 10 nL F0100492A F0100448A J0903005A C0400398A Rainin L-2 H0100116A 0.1-2 nL 2 nL HO100003A Rainin L-5000 D1080497A 0.5-5 mL 5 mL Rainin L-10000 A1058487A 1-10 mL 10 mL Rainin L-20000 L0931886A 2-20 mL 20 mL Percent inaccuracy is recorded from the PCS printout for each pipette. Verification status = PASS if percent inaccuracy is within ±5%. If verification status = FAIL, record "yes" in Corrective Action column and fill in Action Taken below. Corrective Actions: Pipette Serial Number Action Taken 1) 2.) ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 12 of 13 Sample PCS Printout OPP Microbiology Laboratory PI . n 0i< (Jl>ER«TOI? ID: t RE«?EMI LOT «: 3^55? TEMPERATURE CC>: 20 ! ------- SOP No. QC-19-06 Date Revised 01-23-14 Page 13 of 13 Sample Pipette Inventory OPP Microbiology Laboratory MLB Pipette Inrentory Manufacturer Mo-del Xe? Serial Xoi'zL X12523D Gils on Ml® X326-10 BH15232 {Ms on M100 GG05I2T GGC5125 AE 30323 Gils on Distriman BD103I3 OOC45H AG53-243A LD53S33SA C0S239S6A JCT53SS-A C3S25980A Raimn L-1000 C3S2359SA K310U"4A EC1305~"'A 709C233-A JQS0S624A D3303539A E030135-A Manufacturer Model Xo Sirtal Xot z s A0510192A L050S21SA C3S219S2A J075CS05A Rainin L-200 C0S235-2A C3820S61A G0101S59A G31023m J3902S32A CG4Q1654A Rainin L-100 C3S25213A C382523SA A0'53"3S2A CGS220SIA J0'T2-9'32A C0S22312A Ramm L-20 C0S253I5A FOIOC-2SA FGI0044SA 70933035A CG40G3SSA Rainin L-2 KG133116A KC103033A Rainin L-50'30 D13S049M Rainin L-10DM A205S4S""A Rainin L-2000'0 LG931S5SA ------- |