IDEXX Colilert® Test Method for the Simultaneous
Detection of Total Coliforms and E. coli in Water

June 2003

Draft


-------
IDEXX Colilert® Test Method for the Simultaneous
Detection of Total Coliforms and E. coli in Water

1.0	Scope and Application

1.1	This method is intended for use in the simultaneous detection and confirmation of total coliforms
and E. coli in water. Any positive sample for total coliforms is an indication of contamination.
Any positive sample for both total coliforms and E. coli is an acute violation.

1.2	The minimum, non-zero number of bacterial counts detectable with this method is a function of
the dilution scheme used when processing the sample.

1.3	Colilert® method can be applied to fresh waters, drinking waters and waste waters. It can be used
as a Presence/Absence test or quantification with either 5, 10, or 15 tube serial dilution MPN
tubes or with the Quanti-Tray™ system (see package insert) (1).

1.4	Since there can be a wide range of coliform levels in surface waters and wastewaters, dilutions
can be used with this method for detecting and enumerating the actual level.

1.5	Colilert® is not for intended for use with marine waters.

2.0	Summary of Method

2.1	This method is based on Defined Substrate Technology®. This product utilizes nutrient indicators
that produce color/fluorescence when metabolized by total coliforms and E. coli. When the
reagent is added to the sample and incubated, it can detect these bacteria at 1 CFU/100 mL within
24 hours with as many as 2 million heterotrophic bacteria/100 mL present. This test is not
intended for marine waters.

3.0	Definitions

3.1	In this method, coliform bacteria are those bacteria which produce a yellow color, and for E. coli,
also produce a fluorescent signal under a 6-watt, 365-nm UV light after incubation at 35°C ±
0.5°C for 24 hours.

4.0	Interferences

4.1	Heterotrophic bacteria greater than 2,000,000/100 mL can yield a positive reaction for coliforms.
Some water samples containing humic material may have an innate color and a control blank of
the same water sample may be may be required for comparison to the inoculated sample.

5.0	Safety

5.1	The analyst/technician must know and observe the normal safety procedures required in a
microbiology laboratory preparing using and disposing of samples, reagents and materials, and
while operating sterilizing equipment.

5.2	Mouth-pipetting is prohibited.

Draft

1

June 2003


-------
6.0	Equipment and Supplies

6.1	Pipettes, sterile, T.D. bacteriological or Mohr, glass or plastic of appropriate volume.

6.2	Sterile vessels, glass or plastic (free from fluorescence), 100- to 200-mL volume.

6.3	Incubator maintained at 35°C ± 0.5°C.

6.4	6-watt, 365-nm, UV light.

7.0	Reagents

7.1	Sterile deionized or distilled water: Water conforming to specification D1193, reagent water
conforming Type II, Annual Book of ASTM Standards (2). Autoclave at 121°C (15-lb pressure)
for 15 minutes or sterile filter using a 0.22 micron filter into a sterile container.

7.2	Sodium thiosulfate is used to dechlorinate drinking water samples.

7.3	Store Colilert® at 4°C - 30°C away from light. The expiration date is indicated on the package
(12 months from the date of manufacture).

8.0	Sample Collection, Preservation and Storage

8.1	Sampling procedures as described in detail in the USEPA microbiology methods manual, Section
II, A (3) and in Standard Methods for the Examination of Water and Wastewater (4).

8.1.1	Storage Temperature and Handling Conditions: Ice or refrigerate bacteriological samples
at a temperature less than 10°C during transit to the laboratory. Use insulated containers
to assure proper maintenance of storage temperature. Take care that sample vessels are
not totally immersed in water during transit.

8.1.2	Holding Time Limitations: Examine samples as soon as possible after collection. For
drinking water samples do not exceed 30 hours holding time from collection to analysis.
For non-potable water for compliance, do not exceed 6 hours holding time and process
within 2 hours.

9.0	Quality Control

9.1	Quality control should be conducted on each lot of Colilert® or more often as regulations
required. Inoculate 100 mL of sterile water with Quanti-Cult™ or American Type Culture
Collection (ATCC) listed below. Follow the procedure in Section 11.

Quanti-Cult™ Organism

ATCC#

Expected Result

E. coli

25922 or 11775

Yellow, fluorescent

Klebsiella pnuemoniae

31488

Yellow

Pseudomonas aeruginosa

10145 or 27853

Colorless, no fluorescence

Draft

2

June 2003


-------
10.0	Calibration and Standardization

10.1	Check temperatures in incubators daily to insure operation within stated limits.

10.2 Check thermometers at least annually against NIST certified thermometer or one that meets the

requirements of NIST Monograph SP 250-23.

11.0	Procedure

11.1	Presence/Absence

11.1.1	Carefully separate one Snap Pack from the strip taking care not to accidentally open
adjacent pack.

11.1.2	Tap the Snap Pack to ensure that all of the Colilert® powder is in the bottom of the
pack.

11.1.3	Open one pack by snapping back the top at the scoreline.

11.1.4	Add the reagent to the 100-mL water sample contained in a sterile, non-fluorescent
vessel.

11.1.5	Aseptically cap and seal the vessel.

11.1.6	Shake until dissolved.

11.1.7	Incubate for 24 hours at 35°C ± 0.5°C

11.1.8	Read the results at 24 hours. Compare each result against the comparator dispensed
into an identical vessel.

11.1.9	If no yellow color is observed, the test is negative.

11.1.10	If the sample has a yellow color equal to or greater than the comparator, the presence
of total coliforms is confirmed. If color is not uniform, mix by inversion, then recheck.

11.1.11	If the sample is yellow, but lighter than the comparator, it may be incubated an
additional 4 hours (but no more than 28 hours total). If the sample is total coliform
positive, the color will intensify. If it does not intensify, the sample is negative.

11.1.12	If yellow is observed, check vessel for fluorescent by placing a 6-watt, 365-nm UV
light within five inches of the sample in a dark environment. Be sure the light is facing
away from your eyes and towards the vessel. If the fluorescence is greater or equal to
the fluorescence of the comparator, the presence of E. coli is confirmed.

11.2	Quantification

11.2.1	Colilert® can be used for multiple tube MPN analysis (e.g., 5-tube, 10-tube, or 15-
tube serial dilutions). Consult Standard Methods for appropriate MPN tables. For
accuracy and counting range, use the IDEXX Quanti-Tray™ or Quanti-Tray 2000™
and follow the above Presence/Absence Sections 11.1.1 through 11.1.5.

11.2.2	If a dilution is required, use sterile water, not buffered water for making dilutions.
Colilert® is already buffered. Always add Colilert® to the proper volume of diluted
sample after making dilutions.

Draft

3

June 2003


-------
11.2.3	Follow the package insert for Quanti-Tray™ (5). Remove a sterile tray from the
plastic bag and squeeze the tray at the top to open. Pour the sample reagent mixture
from step 11.1.5 above directly into the tray avoiding contact with the foil tab, and
then seal the tray with the Quanti-Tray™ sealer.

11.2.4	Incubate at 24 hours at 35°C ± 0.5°C.

12.0	Data Analysis and Calculations

12.1	Quanti-Tray™

12.1.1	Follow the same interpretation directions from Section 11.1.8 above to count the
number of positive wells. Refer to the MPN table provided with the Quanti-Tray™ to
determine the Most Probable Number (MPN) of total coliforms (yellow wells) and E.
coli (yellow/fluorescent wells) in the sample. The color and fluorescent intensity of
positive wells may vary.

12.1.2	Record the results as the Most Probable Number/100 mL. If any dilutions were made,
multiply the MPN/mL by the dilution factor to obtain the final MPN/100 value.

13.0	Method Performance

13.1	Colilert® found to be equally sensitive to LTB, EC+MUG (6).

13.2	Correlation of 0.905 found between Colilert® and m-Tec for E. coli (6).

14.0	Reporting Results

14.1	Report results as Presence or Absence for total coliforms and for E. coli. For quantification,
report results as MPN/100 mL for total coliforms and E. coli.

15.0	Verification Procedure

15.1	Not applicable.

16.0	Pollution Prevention

16.1	The solutions and reagents used in this method pose little threat to the environment when
recycled and managed properly.

16.2	Solutions and reagents should be prepared in volumes consistent with laboratory use to minimize
the volume of expired materials to be disposed.

17.0	Waste Management

17.1	It is the laboratory's responsibility to comply with all federal, state and local regulations
governing waste management, particularly the biohazard and hazardous identification rules and
land disposal restrictions. Compliance with all sewage discharge permits and regulations is also
required.

17.2	Samples, reference materials and equipment known or suspected to have viable bacteria attached
or contained must be sterilized prior to disposal.

Draft	4	June 2003


-------
18.0	References

18.1	Colilert® Package Insert from IDEXX.

18.2	Annual Book of ASTM Standards, Vol. 11.01, American Society for Testing Materials,
Philadelphia, PA 19103.

18.3	Bordner, R., J.A. Winter and P.V. Scarpino (eds.). Microbiological Methods for Monitoring the
Environment, Water and Wastes, EPA-600/8-78-017. Office of Research and Development,
USEPA.

18.4	Clesceri, L.S., A.E. Greenberg, A.D. Eaton (eds.). 1998. Standard Methods for the Examination
of Water and Wastewater, 20th Edition, American Public Health Association, Washington, DC.

18.5	Quanti-Tray™ Package Insert from IDEXX.

18.6	Federal Register ! Nol. 66, No. 169 / Thursday, August 30, 2001, page 45818.

Draft

5

June 2003


-------