SOP HW-36A Revision 0 June 2015 Hazardous Waste Support Section SOP No. HW-36A Revision 0 SOM02.2 Pesticide Data Validation Approvals: J Russeft Arnone Chemist. Hazardous Waste Support Section C (j /Sc/s Date 2, ArC Cocifzza Chief hazardous Waste Support Section XJai/o Date Jon/Gab^ Chief, Hazardous Waste Support Branch / /? Date ------- SOP IIW-36A Revision 0 June 2015 NOTICE The policies and procedures set forth here are intended as guidance to the United States Environmental Protection Agency (hereafter re 1 erred to as USRPA) and other governmental employees. They do not constitute rule making by I 'Sl'PA, and may not be relied upon to create a substantive or procedural right enforceable by any other person. The Government may take action that is at variance with the policies and procedures in this manual. The guidance for data validation set forth in the quality assurance project plan (QAPP) for the project associated with the data in question will always take precedence over the data validation guidance listed herein. Validators should note that their professional judgment supersedes any guidance listed in this document. Government contractors to the USFPA using this document to validate data should not hesitate to contact their Contracting Officer Representative with any questions regarding data validation or data package completeness. This document can be obtained from the l!STPA's Region 2 Quality Assurance website at: http://www.eDa.gov/reaion2/qa/documents.htm 1 ------- SOP IIW-36A Revision 0 June 2015 tabu; ok com cm s NO TICK I tabu: ok com km s 2 US 1 OI* I AB K K S 3 AC RON V MS ,4 DATA Ql ALIKIKR DKKINITIONS 7 DATA PACKAGK INSPKCHON 7 IIWSS DATA VALIDATION PROCESS 8 PRELIMINARY REVIKW 9 Preservation 10 Gas Chromatograph with Klectron Capture Detector (GC/KCD) Instrument Performance C heck 12 Initial Calibration 15 Continuing Calibration Verification (CCV) 18 Blanks 20 Surrogate Spikes 23 Matrix Spike/Matrix Spike Duplicates (MS/MSDs) 25 Laboratory Control Samples (LCSs) 26 Horisil ( artridge Performance Check 28 Gel Permeation Chromatography (GPC) Performance Check.................................. 29 Target Compound Identification 30 Gas Chromatograph/Mass Spectrometer ((iC/MS) Confirmation 32 Compound Quantitation and Reported Contract Required Quantitation Limits (CRQL.s) 33 Kield Duplicates 34 Overall Assessment of Data 35 APPKNDIX A: GLOSSARY 36 APPKNDIX B: ORGANIC DATA KXECLTIVK NARRATIVE TEMPLATE 39 APPKNDIX C: SAMPLE ORGANIC DATA SAMPLE SUMMARY 40 APPKNDIX I): KLI C I RONIC DATA DELIVERABLE TEMPLATE 41 2 ------- SOP IIW-36A Revision 0 June 2015 LIST OK TABLKS Table 1.1 lolding Time Actions for Pesticide Analyses ......... 11 Table 2. Gas Chromatograph with 1 Electron Capture Detector (GC/KCD) Instrument Performance Check Actions 11 Table 3. Concentration Levels of Calibration Standards.. 15 Table 4. Initial Calibration Actions for Pesticide Analyses 17 Table 5. Continuing Calibration Verification (CCV) Actions for Pesticide Analyses 19 Table 6. Blank Actions for Pesticide Analyses 22 Table 7. Surrogate Actions for Pesticide Analyses 24 Table 10. Pesticides Laboratory Control Sample (I.CS) Spike Compounds and Recovery Limits 26 Table 11. Laboratory Control Sample (LCS) Recovery Actions 27 Table 12. Florisil Cartridge Performance Check Actions 28 Table 13. Gel Permeation Chromatography (GPC) Performance Check Actions 29 Table 14. Action on Qualifying Positive Pesticide Result 31 Table 15. Gas Chromatograph/Mass Spectrometer (GC7IV1S) Conllrmation Actions 32 Table 16. Percent Moisture Actions for Pesticide Analysis For Non-Aqueous Samples 33 3 ------- SOP IIW-36A Revision 0 June 2015 ACRONYMS %l) Percent Difference %RSD Percent Relative Standard Deviation A KG A roc lor ASH Analytical Services Branch BIB Bromofl uoroben/.ene CCS Contract Compliance Screening CCV Continuing Calibration Verification CI Calibration Factor C'LP Contract 1 .aboratory Program CLP PO Contract 1 .aboratory Program Project Officer CRQL Contract Required Quantitation Limit (SI Complete SDG file DART Data Assessment Rapid Transmittal DA 1 Data Assessment Tool DCH Decaehlorobiphenyl 1,1 rp V Decafluorolriphenylphosphine DMC Deuterated Monitoring Compound DQA Data Quality Assessment DQO Data Quality Objective KDI) Electronic Data Deliverable Kl) M EXES Data Manager LSAT Environmental Services Assistance Team IX IS Electronic Data eXchange and Evaluation System GC Gas Chromatograph GC/FXD Gas Chromatograph/Electron Capture Detector GC/MS Gas Chromatograph/Mass Spectrometer GPC Gel Permeation Chromatography nwss 1 la/ardous Waste Support Section IN DA Individual Standard Mixture A INDB Individual Standard Mixture B IN DC Individual Standard Mixture G I.CS Laboratory Control Sample MS Matrix Spike MSI) Matrix Spike Duplicate OSRTI Office of Superfurtd Remediation and Technology Innovation PC lis Polychlorinated Biphenyls pi: Per form ance E v aluation PLM Performance Evaluation Mixture QA Quality Assurance QAC Quality Assurance Coordinator QAPP Quality Assurance Project Plan QC Quality Control RAS Routine Analytical Services 4 ------- SOP IIW-36A Revision 0 June 2015 RIC Reconstructed Ion Chromatogram RIM) Relative Percent Difference RRK Relative Response factor RRF Mean Relative Response f actor RR'I Relative Retention Time RSCC Regional Sample Control Center Coordinator RSI) Relative Standard Deviation RT Retention Time SAP Sampling and Analysis Plan SC P Single Component Pesticide SIM, Sample Delivery Group SIM Selected Ion Monitoring SMO Sample Management OHice SOP Standard Operating Procedure SOW Statement of Work TCL Target Compound List K LP Toxicity Characteristics 1 .eachale Procedure IC X Tetrachloro-m-xylene TIC Tentatively Identified Compound TO PO Task Order Project Officer TR/COC Traffic Report/Chain of Custodv Record US EPA United States Knvironmental Protection Agency liV Ultraviolet VISR Validated Time of Sample Receipt 5 ------- SOP IIW-36A Revision 0 June 2015 INTRODUCTION This document is designed to offer the data reviewer guidance in determining the validity of analytical data generated through the I SI PA Contract Laboratory Program {("I P) Statement of Work (SOW) for Multi-Media, Multi-Concentration Organics Analysis (SOM02.2), and any future editorial revisions of SOM02.2, hereinafter referred to as the SOM02.2 SOW. This guidance is somewhat limited in scope and is intended to be used as an aid in the formal technical review process. The guidelines presented in the document will aid the data reviewer in establishing (a) if data meets the specific technical and QC criteria established in the SOW, and (b) the validity and extent of bias of any data not meeting the specific technical and QC criteria established in the SOW. It must be understood by the reviewer that acceptance of data not meeting technical requirements is based upon many factors, including, but not limited to site-specific technical requirements, the need to facilitate the progress of specific projects, and availability for re- sampling. The reviewer should note that while this document is to be used as an aid in the formal data review process, other sources ofguidance and information, as wel 1 as professional judgment, should also be used to determine the ultimate validity of data, especial 1)' in those cases where all data does not meet specific technical criteria. 6 ------- SOP IIW-36A Revision 0 June 2015 DATA Ql ALII II R DKI INH IONS The following definitions provide brief explanations of the national qualifiers assigned to results in the data review process. u The analyte was analyzed for, but was not detected above the level of the reported sample quantitation limit. J The result is an estimated quantity. The associated numerical value is the approximate concentration of the analyte in the sample. J+ The result is an estimated quantity, but the result may be biased high. J- The result is an estimated quantity, but the result may be biased low. NJ The analysis indicates the presence of an analyte that has been "tentatively identified" and the associated numerical value represents its approximate concentration. LI J The analyte was analyzed for, but was not detected. The reported quantitation limit is approximate and may be inaccurate or imprecise. R The data are unusable. The sample results are rejected due to serious deficiencies in meeting Quality Control (QC) criteria. The analyte may or may not be present in the sample. C T his qualifier applies to results when the identification has been confirmed by Gas Chromatograph/Mass Spectrometer (GC/MS) X This qualifier applies to results when GC/MS analysis was attempted but unsuccessful DATA PACK AG 1*1 INSPECTION For data obtained through the Contract Laboratory Program (CLP), the HXLS Data Manager (LDM) is a useful tool in the data review process. I'or more information about LDM, please refer to the following Sample Management Office (SMO) website: httpsi/zepasmoweb. fedcsc.com/help/uuides/Submit" o2()and°o2()lnspect0 o20Datan o20Quickno20G uide%20°o28LXLS" o29.pdf LDM will identify an}' missing and/or incorrect information in the data package. The CLP laboratory may submit a reconciliation package for any missing items or to correct data. If there are any concerns regarding the data package, contact the CLP Project Oflker (CLP PO) from the Region where the samples were taken. For personnel contact information, please refer to the following CI P website: http://w\vw'.epa.gov/supeiTund/programs/clp/contacts.htrn ------- SOP IIW-36A Revision 0 June 2015 IIVYSS DATA VALIDATION PROCESS Alter downloading the data package from I DM, the data validator will use the recommendations in this SOP as well as their own professional judgment to validate the data. All data is initially marked as "reportable" (Y) in EDM before validation is begun. Sometimes, due to dilutions, re-analyses, or SlM/scan runs all being performed, there will be multiple results for a single analyte from a single sample. The following criteria and professional judgment are used to determine which result should be reported: The analysis with the lower CRQL The analysis with the better QC results The analysis with the higher result The analyte values and their respective CRQI.s are then transferred into a single sample run. The runs that are not to be used are updated as "not reportable" or (N) in KDM. The data will be saved in the following location, under the appropriate case number folder: t.i: DESADIV IIWSS DM A VALIDATION The file naming conventions will consist of A. case number i.e., 12345 B. SDG name i.e., BXYI2 C. level of validation performed i.e., S3VL Examples: 12345JiXYl 2_S3VE.xls 12345 li\YI2 S3VI.M.\Is When data validation is completed, the data package is uploaded for the client to download from the I iWSS data delivery website: The completed data package includes the Executive Narrative (see Appendix B for template), the Sample Summary Report (see Appendix C for example), and the Electronic Data Deliverable (EDD) (see Appendix I) for a list of the column headers included in this document). 8 ------- SOP IIW-36A Revision 0 June 2015 PRKLIMINARY UK VIEW This document is for the review ofanalytical data generated through the SOM02.2 SOW and any future editorial revisions of SOM02.2 for TSTPA Region 2. To use this document effectively, the reviewer should have an understanding of the analytical method and a general overview of the Sample Deliver}' Group (SDG) or sample Case at hand. The exact number of samples, their assigned numbers, their matrix, and the number of laboratories involved in the analysis are essential information. It is suggested that an initial review of the data package be performed, taking into consideration all information specific to the sample data package [e.g.. Modi lied Analysis requests, Traffic Report/Chain of Custody (TR/COC) documentation, SDG Narratives, etc.]. The reviewer should also have a copy of the Quality Assurance Project Plan (QAPP) or similar document for the project for which the samples were analyzed. The criteria for data validation outlined in the QAPP supersede this Standard Operating Procedure. The reviewer should contact the appropriate Regional Contract Laboratory Program Project Officer (C I P PO) to obtain copies of the QAPP and relevant site information. This information is necessary in determining the final usability of the analytical data. The SDGs or Cases routinely have unique samples that require special attention from the reviewer. These include Held blanks and trip blanks. Held duplicates, and Performance Evaluation (PI:) samples which must be identified in the sampling records. The sampling records (e.g.. TR/COC records. Held logs, and/or contractor tables) should identify: 1. The Region where the samples were taken, 2. The Case number, 3. The complete list oI'samples with information on: a. Sample matrix; b. Field blanks (i.e., equipment blanks or rinsate blanks) and trip blanks: c. Field duplicates; d. Field spikes; e. QC audit samples; f. Shipping dates; g. Preservatives; and h. Laboratories involved. The TR/COC documentation includes sample descriptions and date(s) of sampling. The reviewer must consider lag times between sampling and start of analysis when assessing technical sample holding times. T he laboratory's SDG Narrative is another source of general information. Notable problems with matrices, insufficient sample volume for analysis or reanalysis, samples received in broken containers, preservation, and unusual events should be documented in the SDG Narrative. The reviewer should also inspect any email or telephone/communication logs detailing any discussion of sample or analysis issues between the laboratory, the CLP Sample Management Office (SMO). and USHPA Region 2. 9 ------- SOP IIW-36A Revision 0 June 2015 Prcsen ation Action; 1. Qualify aqueous sample results using preservation and technical holding time information as follows (see Table 1): a. If there is no evidence that the samples were properly preserved (T ^ 4CC ± 2°C), and the samples were extracted or analyzed within the technical holding times [seven (7) days from sample collection for extraction; 40 days from sample collection for analysis], qualify detects as estimated (J) and non- detects as estimated (UJ). b. If there is no evidence that the samples were properly preserved (T - 4rC ± 2r'C), and the samples were extracted or analyzed outside the technical holding times [seven (7) days from sample collection for extraction; 40 days from sample collection for analysis], qualify detects as estimated (J) and non- dctects as estimated (UJ). c. If the samples were properly preserved, and were extracted and analyzed within the technical holding times [seven (7) days from sample collection for extraction; 40 days from sample collection for analysis], no qualification of the data is necessary. d. If the samples were properly preserved, and were extracted or analyzed outside the technical holding limes [seven (7) days from sample collection for extraction; 40 days from sample collection for analysis], qua lily detects as estimated (J) and non-detects as estimated (UJ). Note in the Data Review Narrath e that holding times were exceeded and the effect of exceeding the holding time on the resulting data, 2. Qualify non-aqueous sample results using preservation and technical holding time information as follows (see Table 1): a. If there is no evidence that the samples were properly preserved (T - 4CC ^ 2CC), and the samples were extracted or analyzed within the technical holding time [14 days from sample collection for extraction; 40 days from sample collection for analysis], qualify detects as estimated (J) and non- detects as estimated (UJ). b. If there is no evidence that the samples were properly preserved (T - 4 C = 2CC), and the samples were extracted or analyzed outside the technical holding time [14 days from sample collection for extraction; 40 days from sample collection for analysis], qualify detects as estimated (J) and non- detects as estimated (UJ). c. If the samples were properly preserved, and were extracted and analyzed within the technical holding time [14 days from sample collection for extraction; 40 days from sample collection for analysis], no qualification of the data is necessary. d. If the samples were properly preserved, and were extracted or analyzed outside the technical holding time 114 days from sample collection for 10 ------- SOP IIW-36A Revision 0 June 2015 extraction; 40 days from sample collection for analysis], qualify detects as estimated (J) and nun-detects as estimated (UJ). Note in the Data Review Narrative that holding times were exceeded and the effect of exceeding the holding time on the resulting data. 3. Whenever possible, the reviewer should comment on the effect of the holding time exceedance on the resulting data in the Data Rev iew Narrative. 4. Use professional judgment to qualify samples whose temperature upon receipt at the laboratory is either below 2 degrees centigrade or above 6 degrees centigrade, 5. If technical holding times are grossly exceeded, use professional judgment to qualify the data. 6. Note, for Contract Laboratory Program Project Officer (CLP PO) action, when technical holding times are exceeded. Table I. Holding Time Actions for Pesticide Analyses Matrix Preserved Criteria Action Detected Associated Compounds Non-Detected Associated Compounds Aqueous No < 7 days (for extraction) < 40 days (for analysis) J UJ Nu > 7 days (for extraction) > 40 days (for analysis) J I 1 Yes 1 7 days {for extraction) < 40 days (for analysis) No qualification Yes > 7 days (for extraction) > 40 days (for analysis) J UJ Yes/No Grosslv F.xceeded Use professional judgment Non-Aqueous No £ 14 days (for extraction) < 40 days (for analysis) J UJ No > 14 days (for extraction) > 40 days (for analysis) J UJ \ es < 14 days (for extraction) i 40 days (for analysis) No qualification Yes ¦> 14 days (for extraction) > 40 days (for analysis) J UJ Yes No Grossh I Acceded Use professional judgment ------- SOP IIW-36A Revision 0 June 2015 Gas Clirom;uograi)h with Electron Capture Detector ((rC/'ECl)) Instrument Performance Check Action: 1. Resolution Check Mixture a. The resolution between two adjacent peaks in the Resolution Check Mixture must be greater than or equal to 80.0% for all analytes for the primary column and greater than or equal to 50.0° o lor the confirmation column in order to use one Individual Standard Mixture (C). The resolution between two adjacent peaks in the Resolution Check Mixture must be greater than or equal to 60.0% if two Individual Standard Mixtures (A and B) are to be used, irresolution criteria are not met, the quantitative results may not be accurate due to inadequate resolution. Qualitative identifications may also be questionable if coelution exists. i. Qualify detects lor target compounds that were not adequate!} resolved as tentatively identilied (NJ) (see Table 2). ii. Qualify non-detected compounds as unusable (R). 2. PHM a. If PHM analysis is not performed at the required frequency (see Pesticides Organic Analysis, of SOM02.2 NFG), qualify all associated sample and blank results as unusable (R). b. The resolution between any two adjacent peaks in the initial calibration and continuing calibration verification PHMs must be greater than or equal to 90% on each GC column. If PHM resolution criteria are not met. the quantitative results may not be accurate due to inadequate resolution. Qualitative identifications may be questionable if coelution exists. i. Qualify detects as tentatively identified (NJ). ii. Qualify non-del eels as unusable (R). c. If4,4'-DDT breakdow n is greater than 20.0%: i. Qualify detects for 4.4'-l)I) f as estimated (J). ii. Qualify detects for 4,4'-DDD and/or 4,4'-DI)H as estimated (J). iii. If4,4'-I)D H was not detected, but 4.4'-I)DD and/or 4.4'-l)I)H were detected, qualify non-detects for 4.4'-I)I)T as unusable (R). and qualify delects for 4.4'-!)!)!) and/or 4,4'-DDH as presumptively present at an approximated quantity (NJ). d. 1 f Hndrin breakdown is greater than 20.0%: i. Qualify detects for Hndrin as estimated (J). ii. Qualify detects for Hndrin aldehyde and/or Hndrin ketone as estimated (J). iii. If Hndrin was not detected, but Hndrin aldehyde and/or Hndrin ketone were detected, qualify the non-detects for Hndrin as unusable (R), and qualify detects for Hndrin aldehyde and/or Hndrin ketone as presumptively present at an approximated quantity (NJ). I _ ------- SOP IIW-36A Revision 0 June 2015 e. If the combined 4,4'-DI)T and Kndrin breakdown is greater than 30.0%. the review er should consider the degree of individual breakdown of4„4'-DDT and Kndrin and apply qualifiers as described in this section. Mid-point Individual Standard Mixtures (A and R) or (C) a. If mid-point Individual Standard Mixture analysis is not performed at the required frequency (see Pesticides Organic Analysis, of SOM02.2 NFG), quality all associated sample and blank results as unusable (R). b. The resolution between any two adjacent peaks in the mid-point concentration of Individual Standard Mixtures (A and B) in the initial calibration and continuing calibration verification must be greater than or equal to 90.0% on each column. The resolution between any two adjacent peaks in the mid-point concentration of Individual Standard Mixture (C) in the initial calibration and continuing calibration verification must be greater than or equal to 80.0°6 for the primary column and greater than or equal to 50.0% for the secondary column. If mid-point Individual Standard Mixtures (A and B) or (C) resolution criteria are not met, the quantitative results may not be accurate due to inadequate resolution. Qualitative identifications may be questionable if coelution exists. i. Qualify detected target compounds that were not adequately resolved tentatively identified (NJ). ii. Qualify non-detecls as unusable (R). Note in the Data Review Narrative the potential efleets on the sample data resulting from the instrument performance check criteria. Notify the Contract Laboratory Program Project Officer (CLP PO) if the laboratory has repeatedly lai led to comply with the requirements for linearity, resolution, or 4,4'-DD 1'I 'ndrin breakdown. ------- SOP IIW-36A Revision 0 June 2015 Table 2. Gas Chromatograph with Klectron Capture Detector (G( 7K( 1>) Instrument Performance Check Actions Criteria |Individual Standard Mixtures ( A and li)| Criteria (Individual Standard Mixture C) Action Resolution Check Mixture °0 Resolution <60.0% Resolution Check Mixture % Resolution <80.0% (primary column) % Resolution <50.0% (secondary column) Detects; Ml Non-detects: R PKM °o Resolution <90.0% Detects: NJ Non-detects: R PKM: 4,4'-DDT °o Breakdown >20.0% and 4,4'-DDT is detected Detects for 4.4'-DI)T: J Detects for 4,4'-DDD: .1 Detects for 4,4'-DDI:: J PKM: 4.4'-DDT % Breakdown -'20.0% and 4.4'-f)I)i is not detected Non-detects for 4.4'- DDT: R Detects for 4.4'-I)I)l): NJ Detects for 4.4'-DDI-:: NJ PEM: Kndrin % Breakdown >20.0% and Kndrin is detected Detects for Tndrin: J Detects for Kndrin aldehy de: J Detects for Kndrin ketone: J PKM: Kndrin % Breakdown >20.0% and Kndrin is not detected Non-detects for Kndrin: R Delects for Kndrin aldehyde: NJ Detects for Kndrin ketone: NJ PKM: Combined °o Breakdown >30°o Apply qualifiers as described above considering degree of individual breakdown. Mid-point Individual Standard Mixtures (A and B) % Resolution <90.0% Mid-point Individual Standard Mixture (C) % Resolution <80.0% (primary column) Mid-point Individual Standard Mixture (C) % Resolution <50.0% (secondary column) Detects: NJ Non-detects: R PKM analysis not performed at the required Frequency (see Pesticides", ofSOM02.2 NFG) All results: R Mid-point Individual Standard Mixtures analysis not performed at the required frequency (see Pesticides, of SOM02.2 NFG) All results: R 14 ------- SOP IIW-36A Revision 0 June 2015 Inilial Calibration Table3, Concentration Levels of Calibration Siaiulards ( (impound Concern ration (n«/mL) CS1 CS2 CS3 CS4 CSS alpha-BHC 5.0 10 20 40 80 gamma-BHC 5.0 10 20 40 80 Heptachlor 5.0 10 20 40 80 Hndosulfan I 5.0 10 20 40 80 Dieldrin 10 20 40 80 160 Hndrin 10 20 40 80 160 4.4-1)1)1) 10 20 40 80 160 4.4'-!)l) 1 10 20 40 80 160 Methoxvchlor 50 100 200 400 800 beta-BI-IC 5,0 10 20 40 80 delta-BHC 5.0 10 20 40 80 Aldrin 5.0 10 20 40 80 1 leptachlor-epoxide 5.0 10 20 40 80 4,4'-DDH 10 20 40 80 160 Hndosullan 11 10 20 40 80 160 Hndosulfan sulfate 10 20 40 80 160 Hndrin ketone 10 20 40 80 160 Hndrin aldehvde 10 20 40 80 160 alpha-Chlordane 5.0 10 20 40 80 gamma-C'hlordane 5.0 10 20 40 80 Tetrachloro-m-xvlene 5.0 10 20 40 80 Decaehlorobiphenyl 10 20 40 80 160 Toxaphene 500 1000 2000 4000 8000 Action,: NOTES: At least one chromatogram from each of the Individual Standard Mixtures (A and B) or (C) must yield peaks that give recorder defections between 50-100% oi"full scale. Hither peak area or peak height may be used to calculate the Calibration Factors (CTs) that are, in turn, used to calculate °oRSl). However, the type of peak measurement used to calculate each Cl; lor a given compound must be consistent. For example, if peak area is used to calculate the low-point CF lor Hndrin,, the mid-point and high-point CFs for Hndrin must also be calculated using peak area. 1. If the proper initial calibration sequence is not performed, or the steps of the initial calibration are not followed in the proper sequence, use professional judgment to evaluate 1^ ------- SOP IIW-36A Revision 0 June 2015 the effect on the data and notify the Contract Laboratory Program Project Officer (CLP PO) (see Table 4). This is especially critical for the low-level standards and non-detects. 2. If RT Windows are not calculated correctly, recalculate the windows and use the corrected values for all evaluations. 3. If the chroniatogram display (recorder deflection) criteria are not met, use professional judgment to evaluate the effect on the data. 4. If the standard concentrations listed in Table 3 are not used, use professional j udgment to evaluate the effect on the data and notify the CLP PO. This is especially critical for the low-level standards and non-detects. 5. The Percent Relative Standard Deviation (%RSD) of the CPs for each of the single component target compounds must be less than or equal to 20.0%, except for alpha-BHC and delta-BI IC. The %RSD of the CFs for alpha-BHC and delta-BHC must be less than or equal to 25.0° o. The %RSD of the CFs for each of the Toxaphene peaks must be £ 30% when 5-point ICAL is performed.The %RSD of the CFs for the two surrogates (tetrachloro-m-xylene and decachlorobipheny 1) must be less than or equal to 30.0%. If the %RSD criteria are not met, qualify detects as estimated (J) and use professional judgment to qualify non-detected target compounds. 6. If the %RSD criteria are within allowable limits, no qualification of the data is necessary. 7. At the reviewer's discretion, and based on the project-specific data quality objectives, consider a more in-depth review using the following guidelines: a. If any pesticide target compound has a %RSI) greater than the maximum criterion, and if eliminating either the high or the low-point of the curve does not restore the %RSI) to less than or equal to the required maximum: i. Qualify detects for that compound(s) as estimated (J). ii. Qualify non-detected pesticide target compounds using professional judgment, b. If the high-point of the curve is outside of the linearity criteria (e.g.. due to saturation): i. No qualifiers are required for detects in the linear portion of the curve. ii. Qualify detects outside oflhe linear portion of the curve as estimated (J). iii. No qualifiers are required for pesticide target compounds that were not detected. c. If the low-point oflhe curve is outside of the linearity criteria: i. No qualifiers are required for detects in the linear portion oflhe curve. ii. Qualify low-level detects in the area of non-linearity as estimated (J). iii. For non-detected pesticide compounds, use the lowest point of the linear portion of the curve to determine the new quantitation limit. 8. Note in the Data Review Narrative potential effects on the sample data due to problems with calibration. Notily the CLP PO if the laboratory has repeated!)' failed to comply with the requirements for frequency, linearity, RT. or resolution. 9. Quality data for Toxaphene sharing the same RT Window with any Single Component Pesticide (SCP) in any Individual Standard Mixture using professional judgment. 10 ------- SOP IIW-36A Revision 0 June 2015 Tahle4. Initial Calihralion Actions for Pesticide Analyses Criteria Action Detected Associated Compounds Non-Detected Associated Com pounds Initial calibration is not performed or not performed in the proper sequence Use professional judgment and notify CLP PO °oRSD exceeds allowable limits* J Use professional judgment 0c»RSI) within allowable limits* No qualification * %RSD 2_20.0°'o for single component target compounds except alpha-BIlC and delta-B!!('. {, oRSI) <25.0% for alpha-Bl IC and delta-BUC. %RS1) <30.0? o for Toxaphene peaks. " uRSI) <_30.0% for surrogates (letrachloro-m-xvlene and decachlorobiphenyl). 17 ------- SOP IIW-36A Revision 0 June 2015 Continuing Calibration Verification (C'C'V) Action; 1. The RT Windows are used in qualitative identilleation. If the standards do not fall within the R I Windows, carefully evaluate the associated sample results (see Table 5). All samples injected alter the last in-control standard are potentially affected. a. For non-detected target compounds in the affected samples, check to see if the sample chromatograms contain any peaks that are close to the expected RT Window of the pesticide of interest. i. If no peaks are present, consider non-detected values to be valid and no qualification of the data is necessary. ii. If any peaks are present close to the expected R I Window of the pesticide of interest, use professional judgment to qualify the non-detects as presumptively present fNJ). b. For detected compounds in the affected samples, if the peaks are within the RT Window, no qualification of the data is necessary. However, if the peaks are close to the expected RT Window of the pesticide of interest, the reviewer may take additional effort to determine if sample peaks represent the compounds of interest. For example, the reviewer can examine the data package for the presence of three or more standards containing the pesticide of interest that were run within the analytical sequence during which the sample was analyzed. If three or more such standards are present, the RT Window can be re-evaluated using the Mean Retention Times (RTs) of the standards. i. If the peaks in the affected sample fall within the revised window, qualify detects as tentatively identified (NJ). ii. If the reviewer cannot do anything with the data to resolve the problem of concern, qualify all non-detects as unusable (R). 2. For the P1:M, if the Percent Difference is not within ±25.0° o, qualify associated detects as estimated (J) and non-detects as estimated (UJ). 3. For the Calibration Verification Standard (CS3), if the Percent Difference is not within ±25.0°o, qualify associated detects as estimated (J» and non-detects as estimated (UJ). 4. If 4,4'-DDT 0..Breakdown is - 20.0%, qualify detected 4,4'-DDT, 4,4'-DDD, and 4,4'-DDE as estimated (J). When 4.4-DDT is not detected, but 4,4'-DDD and 4.4'-DDH are detected, qualify non-detected 4,4'-DDT as unusable (R) and detected 4,4'-DDD and 4,4 -1)1)1. as presumptively present with estimated concentration (NJ). If Endrin %Breakdo\vn > 20.0%. qualify detected Endrin, Endrin aldehjde and Hndrin ketone as estimated (J). When Endrin is not detected, but Endrin aldehyde and Endrin ketone are detected, 5. qualify non-detected Endrin as unusable (R) and detected Endrin aldehyde and Endrin ketone as presumptively present with estimated concentration (NJ). If more than 14 hours has elapsed from the injection of the instrument blank that begins an analytical sequence (opening (TV) and the injection of either a PI: VI or mid-point concentration of the Individual Standard Mixtures (A and B) or (C). qualify all data as unusable (R). 18 ------- SOP IIW-36A Revision 0 June 2015 6. 11" more than 12 hours has elapsed from the injection of the instrument blank that begins an analytical sequence (opening CCV) and the injection of the last sample or blank that is part of the same analytical sequence, qualify all data as unusable (R). 7. H" more than 72 hours has elapsed from the injection of the sample with a Toxaphene detection and the Toxaphene Calibration Verilocation Standard (CS3), qualify all data as unusable (R). 8. If the Percent Difference, time elapsed, and R Is are within acceptable limits, no qualification of the data is necessary. 9. Note in the Data Review Narrative potential effects on the sample data due to problems with calibration. Fable 5. Continuing Calibration Verification (CCV) Actions for Pesticide Anahses Criteria Action Detected Associated Compounds Non-Detected Associated Compounds RT outofRT window Use professional judgment °oD not within limits* J rj Time elapsed is greater than acceptable limits** R °oD, time elapsed, and R f are within acceptable limits No qualification PEM: 4.4'-DDT ^.Breakdown >20.0% and 4.4'- DDT is detected j for 4,4'-DDT. 4,4'- 1)1)1). and 4.4'-l)l)E No qualification PKM: 4,4'-I)DT % Breakdown --20.0% and 4,4'- I)D I is not detected R lor 4.4 -1 )[> I \.I lor 4.4-1)1)1) and 4,4'-f)f)i; PKM: Endrin %Breakdo\\n -"20.0% and Hndrin is detected j for Endrin, Endrin aldehyde, Endrin ketone No qualification PEM: Endrin "'((Breakdown ""--20.0% and Endrin is not detected R for Endrin NJ for aldehyde and Endrin ketone See Actions 2 and 3 See Actions 5, 6, and 7 19 ------- SOP IIW-36A Revision 0 June 2015 Blanks Action: NOTES: The concentration of non-target compounds in all blanks must be less than or equal to 10 j.ig/1,. The concentration of each target compound found in the method or Held blanks must be less than its CRQL listed in the method. Data concerning the Held blanks are not evaluated as part of the CCS process. If field blanks are present, the data reviewer should evaluate this data in a similar fashion as the method blanks. NOTES: "Water blanks, "drill blanks", and "distilled water blanks" are validated like any other sample and are not used to qualify data. Do not confuse them with the other QC blanks discussed below. All Held blank results associated with a particular group of samples (may exceed one per case) must be used to qualify data. Blanks may not be qualified because of contamination in another blank. Field blanks must be qualiiied for system monitoring compounds, instrument performance criteria, and spectral or calibration QC problems. Analytes qualiiied "U" for blank contamination are treated as "hits" when qualifying for calibration criteria. Samples taken from a drinking water tap do not have associated field blanks. When applied as described in Table 6 below, the contaminant concentration in the blank is multiplied by the sample dilution factor. Action regarding unsuitable blank results depends on the circumstances and origin of the blank. In instances where more than one of the same type of blank is associated with a given sample, qualification should be based upon a comparison with the associated blank having the highest concentration of a contaminant. Do not correct the results by subtracting any blank value. 1. If a target SCP or Toxaphene is found in a method blank, but not found in the sample, no qualification of the data is necessary (see 'fable 6). 2. If a target SCP or Toxaphene concentration in a method or Held blank is less than the CRQI. and: a. the sample concentration is less than the CRQI.. report the CRQI. value with a "IT. b. the sample concentration is greater than or equal to the CRQL, no qualification is required. 3. If a target SCP or Toxaphene concentration in a method or Held blank is greater than the CRQL mid: a. the sample concentration is less than the CRQI., report the CRQL value with a "ir. b. the sample concentration is greater than or equal to the CRQL, and less than or equal to the blank concentration, report the concentration of the compound in the sample at the same concentration found in the blank and qualify with a "IP, 20 ------- SOP IIW-36A Revision 0 June 2015 c. the sample concentration is greater than or equal to the CRQL and greater than the blank concentration, no qualification is required. 4. if a target SCT or Toxaphene concentration in a method or field blank is equal to the CRQLand: a. the sample concentration is less than or equal to the CRQL, report the CRQL value with a "lr, b. the sample concentration is greater than the CRQL, no qualification is required. 5. If gross contamination exists (i.e., saturated peaks, "hunip-o-grams", "junk" peaks), raise the CRQL to the level of the blank contamination and report the associated sample data below this level as CRQL-U, Non-detected pesticide target compounds do not require qualification unless the contamination is so high that it interferes with the analyses of non-detected compounds. 6. If contaminants are found in the field blanks, the following is recommended: a. Review the associated method blank data to determine if the contain inant(s) was also present in the method blank. i. If the analyte was present at a comparable level in the method blank, the source of the contamination may be in the analytical system and the action recommended for the method blank would apply. ii. If the analyte was not present in the method blank, the source of contamination may be in the storage area, in the field, or during sample transport. Consider all associated samples for possible cross- contamination. 7. If method blank data are unavailable, the reviewer may use professional judgment or substitute field blank data for missing method blank data. NOTE: There may be instances where little or no contamination was present in the associated blanks, but qualification of the sample is deemed necessary. If the reviewer determines that the contamination is from a source other than the sample, they should qua lily the data. Contamination introduced through dilution water is one example. Although it is not always possible to determine, instances of this occurring can be detected when contaminants are found in the diluted sample result, but are absent in the undiluted sample result. 21 ------- SOP IIW-36A Revision 0 June 2015 Table 6. Blank Actions for Pesticii e Analyses Blank Type Blank Result Sample Result Action for Samples Method. Sulfur Cleanup, Instrument Field, TCLP/SPLP Detects Not delected No qualilication required - C'RQI. C'RQI. Report C'RQI. \alue with a U C'RQi. No qualification required C'RQI. C'RQl. Report C RQI. value with a U C'RQI. and . blank concentration Report blank value lbr sample concentration w ith a t" C'RQi and blank concentration No qualilication required C'RQI. IA r O Report C'RQI value with a U C'RQI. No qualilication required Gross contamination Detects Report blank value lor sample concentration with a U ------- SOP IIW-36A Revision 0 June 2015 Surrogate Spikes Action: If either surrogate spike recover} is outside the acceptance limits, the reviewer must consider the existence ofcoelution and interference in the raw data and use professional judgment to qualify data, as surrogate recovery problems may not directly apply to target analytes. 1. For any surrogate recover)1 greater than 150% (see Table 7): a. Qualify detected target compounds as biased high (J+). b. Do not quality non-detected target compounds. 2. If both surrogate recoveries are greater than or equal to 30%, and less than or equal to 150%, no qualification of the data is necessary. 3. For any surrogate recovery greater than or equal to 10%. and less than 30%: a. Qualify detected target compounds as biased low (J-), b. Qualify non-detected target compounds as approximated (UJ). 4. For any surrogate recover}1 less than 10%, the reviewer should examine the sample chromatogram to assess the qualitative validity of the analysis. If low surrogate recoveries are from sample dilution, professional judgment should be used to determine if the resulting data should be qualified. If sample dilution is not a factor: a. Qualify detected target compounds as biased low (J-), b. Quali ly non-detected target compounds as unusable (R). 5. In the special case of a blank analysis with surrogates out of specification, the reviewer must give special consideration to the validity of associated sample data. The basic concern is whether the blank problems represent an isolated problem with the blank alone, or whether there is a fundamental problem with the analytical process. For example, if one or more samples in the batch show acceptable surrogate recoveries, the reviewer may choose to consider the blank problem to be an isolated occurrence. Note, for Contract I.aboratory Program Project Ofticer (CLP PO) action, analytical problems even if this judgment allows some use of the affected data. 6. If surrogate RTs in PI {Ms, Individual Standard Mixtures, samples, and blanks are outside of the R I' Windows, the reviewer must use professional judgment to qualify data. 7. If surrogate RTs are within R I windows, no qualification of the data is necessary. 8. If the two surrogates were not added to all samples, MS/MSDs, standards, FCSs, and blanks, use professional judgment in qualifying data as missing surrogate analyte ma}1 not directly apply to target analytes. 23 ------- SOP IIW-36A Revision 0 June 2015 Table 7. Surrogate Actions lor Pesticide Analyses Action* Criteria Detected Target Non-detected Target Com pounds Compounds %R> 150% J+ No qualillcation 30% < %R < 150% No qualification 10°o < °0R < 30°o J- UJ %R < 10% (sample dilution not a factor) J- R °oR < 10% (sample dilution is a factor) Use professional judgment RT out of RT window Use professional judgment RT within R 1 window No qualification Use professional judgment in qualifying data, as surrogate recovery problems may not directly apph to target analytes. 24 ------- SOP IIW-36A Revision 0 June 2015 Matrix Spike/Matrix Spike Duplicates (MS/MSI)s) Action; NOTES: Data lor MS and MSDs will not be present unless requested by the Region. Notify the Contract Laboratory Program Project Officer (CLP PO) if a Held blank was used for the MS and MSI), unless designated as such by the Region. NO IT.: Lor a Matrix Spike that does not meet criteria, apply the action to only the field sample used to prepare the Matrix Spike sample. If it is clearly stated in the data validation materials that the samples were taken through incremental sampling or some other method guaranteeing the homogeneity of the sample group, then the entire sample group may be qualified. 1. No qualification of the data is necessary on MS and MSI) data alone. However, using professional judgment, the validator may use the MS and MSI) results in conjunction with other QC criteria and determine the need for some qualification of the data. 25 ------- SOP IIW-36A Revision 0 June 2015 Laboratory Control Samples 1 LCSs) Table 10. Pesticides Laboratory Control Sample (LCS) Spike Compounds and Recovery Limits LCS Spike Compound Recovery Limits (%) gamma-BI IC 50 - 120 Heptachlor epoxide 50 - 150 Dieldrin 30-130 4,4"-DDl: 50 - 150 Lndrin 50 120 F.ndosullan sulfate ¦ji 0 1 to o trans-Chlordane 30 - 130 I etrachloro-m-xy 1 ene (surrogate) 30 - 150 Deeachlorobiphenyl (surrogate) 30 - 150 Action: NOTES: The recovery limits for any of the compounds in the l.CS may be expanded at any- time during the period of performance if USLPA determines that the limits are too restrictive. All samples prepared and analyzed with an LCS that does not meet the technical acceptance criteria in the method will require re-extraction and re-analysis. If the ICS criteria are not met, laboratory performance and method accuracy are in question. Use professional judgment to determine if the data should be qualified or rejected. The following guidance is suggested for qualifying sample data for which the associated LCS does not meet the required criteria. 1. If the LCS recover}' criteria are not met, use the 1 .CS results to qualify sample data for the specific compounds that are included in the ICS solution (see Table 11). a. If the LCS recovery exceeds the upper acceptance limit, qualify detected target compounds as estimated (J). Do not qualify non-detected target compounds. b. If the LCS recovery is less than the lower acceptance limit, qualify detected target compounds as estimated (J) and non-detects as unusable (R). c. Use professional judgment to qualify data for compounds other than those compounds that are included in the l.CS. d. l-se professional judgment to qualify non-1 .CS compounds. Take into account the compound class, compound recovery efficiency, analytical problems associated with each compound, and comparability in the performance of the LCS compound to the non-LCS compound. 2. If the LCS recovers- is within allowable limits, no qualification of the data is necessary. 3. Note, for Contract Laboratory Program Project Officer (CLP PO) action, if a laboratory' fails to analyze an LCS with each Sample Delivery Group (SDG), or if the reviewer has knowledge that a laboratory consistently fails to generate acceptable LCS recoveries. 26 ------- SOP IIW-36A Revision 0 June 2015 Tahle 11. Laboratory Control Sample (LCS) Recovery Actions Action Criteria Detected Non-Detected Associated Associated Compounds Com pounds °oR > Upper Acceptance Limit J+ No qualification °oR < Lower Acceptance Limit J- R Lower Acceptance Limit < °oR < Upper Acceptance Limit No qualification ------- SOP IIW-36A Revision 0 June 2015 Florisil Cartridge Performance ( heck Action; NOTE: Florisil cartridge cleanup is mandatory for aU extracts. 1. If the Florisil Cartridge Performance Check criteria are not met, examine the raw data for the presence of polar interferences and use professional judgment in qualifying the data as follows (see Table 12): a. I f the Percent Recovery is greater than 120% for any of the pesticide target compounds in the Florisil Cartridge Performance Check, qualify detected compounds as estimated (J). Do not quality non-detected target compounds. b. If the Percent Recover)1 is greater than or equal to 80% and less than or equal to 120% for all the pesticide target compounds, no qualification of the data is necessary. c. If the Percent Recover}' is greater than or equal to 10% and less than 80% for any of the pesticide target compounds in the Florisil Cartridge Performance Check, qualify detected target compounds as estimated (J) and non-detected target compounds as approximated (UJ). d. If the Percent Recovery is less than 10% for any of the pesticide target compounds in the Florisil Cartridge Performance Check, qualify detected compounds as estimated (J) and qualify non-detected target compounds as unusable (R). e. If the Percent Recovery of 2,4,5-trichlorophenol in the Florisil Cartridge Performance Check is greater than or equal to 5%. use professional judgment to qualify detected and non-detected target compounds, considering interference on the sample chromatogram. 2. Note in the Data Review Narrative potential effects on the sample data resulting from the Florisil Cartridge Performance Check analysis not yielding acceptable results. Table 12. Morisil Cartridge Perfor mance Check Actions Action Criteria Detected Associated Non-Detected Compounds Associated Compounds %R > 120% (pesticide target compounds) J No qualification 80% < °oR < 120% No quali ication 10% < %R -'80°o (pesticide target j UJ compounds) %R < 10% (pesticide target compounds) J R %R > 5% (2,4,5-trichlorophenol) Use professional judgment* Check sample chromatogram for interferences. ------- SOP IIW-36A Revision 0 June 2015 (.el Permeation Chromatography (GPC) Performance Check Action; NOTE: GPC cleanup is mandatory for all soil samples. 1. If GPC criteria are not met, examine the raw data for the presence of high molecular weight contaminants; examine subsequent sample data for unusual peaks; and use professional judgment in qualilying the data. Notify the Contract Laboratory Program Project Officer (CLP PO) if the laboratory chooses to analyze samples under unacceptable GPC criteria. 2. If the Percent Recovery is less than 10% for the pesticide compounds and surrogates during the GPC calibration check, the non-detected target compounds may be suspect. Qualify detected compounds as estimated (J) (see 'fable 13). Qualify all non-detected target compounds as unusable (R). 3. If the Percent Recover}' is greater than or equal to 10% and is less than 80% for any of the pesticide target compounds in the GPC calibration, qualify detected target compounds as estimated (J) and non-detected target compounds as approximated (L"J). 4. If the Percent Recovery is greater than or equal to 80% and less than or equal to 120% for all the pesticide target compounds, no qualification of the data is necessary. 5. If high recoveries (i.e., greater than 120%) were obtained for the pesticides and surrogates during the GPC calibration check, qua lily detected compounds as estimated (J). Do not qualify non-detected target compounds. 6. Note in the Data Review Narrative potential effects on the sample data resulting from the GPC cleanup analyses not y ielding acceptable results. Table 13, Gel Permeation Chromatography (GPC) Performance Cheek Actions Action Criteria Detected Associated Non-Detected C om pounds Associated Com pounds %R 10% (pesticide target compounds) J R 10%-%R -- 80% J UJ 80%i£%R^ 120% No quali 1 cat ion %R 120°o (pesticide target compounds) J No qualilication 29 ------- SOP IIW-36A Revision 0 June 2015 In met Coin pound Identification Criteria: 1. The Retention Times (RTs) of both of the surrogates and reported target compounds in each sample must be within the calculated RT Windows on both columns. Tetrachloro- m-xylene (TCX) must be within ±0,05 minutes of the Mean RT ( RT ) determined from the initial calibration and Decachlorobiphenyl (DCB) must be within ±0.10 minutes of the RT determined from the initial calibration. 2. The Percent Difference (°oD) for the detected mean concentrations of a pesticide target compound between the two Gas Chromatograph (GC) columns must be within the inclusive range of ^25.0. 3. When no analytes are identified in a sample, the chromatograms from the analyses of the sample extract must use the same scaling factor as was used for the low-point standard of the initial calibration associated with those analyses. 4. Chromatograms must display Single Component Pesticides (SCPs) detected in the sample and the largest peak of any multi-component analyte detected in the sample at less than full scale. 5. If an extract must be diluted, chromatograms must display SCPs peaks between 10-100% of full scale, and multi-component analytes between 25-100% of full scale. 6. For an>' sample, the baseline of the chromatogram must return to below 50% of full scale before the elution time of alpha-BHC, and also return to below 25% of lull scale after the elution lime of alpha-BHC and before the elution time of DCB. 7. If a chromatogram is replotted electronically to meet these requirements, the scaling factor used must be displayed on the chromatogram, and both the initial chromatogram and the replotted chromatogram must be submitted in the data package. Action: 1. If the qualitative criteria for both columns were not met, all target compounds that are reported as detected should be considered non-detected. The reviewer should use professional judgment to assign an appropriate quantitation limit using the following guidance: a. If the detected target compound peak was sufficiently outside the pesticide RT Window, the reported values may be a false positive and should be replaced with the sample Contract Required Quantitation Limits (CRQL) value. b. If the detected target compound peak poses an interference with potential detection of another target peak, the reported value should be considered and qualified as unusable (R). 2. If the data reviewer identifies a peak in both GC column analyses that falls within the appropriate RT Windows, but was reported as a non-detect, the compound may be a false negative. Use professional judgment to decide if the compound 30 ------- SOP IIW-36A Revision 0 June 2015 should be included. Note in the Data Review Narrative all conclusions made regarding target compound identification. 3. If the Toxaphene peak RT windows determined from the calibration overlap with NCI's or chromatographic interferences, use professional judgment to qualify the data. 4 If target compounds were detected on both GC columns, and the Percent Difference between the two results is greater than 25.0%, consider the potential for coelution and use professional judgment to decide whether a much larger concentration obtained on one column versus the other indicates the presence of an interfering compound. 1 f an interfering compound is indicated, use professional judgment to determine how best to report, and if necessary, qualify the data according to the guidelines in Table 14 below. 5. If Toxaphene exhibits a marginal pattern-matching quality, use professional judgment to establish whether the differences are due to environmental "weathering" (i.e., degradation of the earlier eluting peaks relative to the later eluting peaks). If the presence of Toxaphene is strongly suggested, report results as presumptively present (N). Table 14, Action on Qunlit'ving iVsithe Posticicle Result Percent Differences Qualifier 0°o - 25°o No qualification 26° o - 200% Professional judgment 1010o — 200% (interferences detected)* JN - 50°o (pesticide value < CRQL)** U >200% R When interferences are detected on either column, qualify the data as "JN". When the pesticide value is below CRQL and °ol) > 50%. raise the value to CRQL and qualify "IT undetected. ------- SOP IIW-36A Revision 0 June 2015 Gas Chromalograph/Mass Spec!rometer ((.C'/iVIS) Confirmation Action; NOTE: This confirmation is not usually provided by the laboratory. In cases where it is provided, use professional judgment to determine if data qualified with "*C can be salvaged if it was previously qualified as unusable (R). 1. If the quantitative criteria for both columns were met (> 5.0 ng/pf for SCPs and > 125 tig/pf, for Toxaphene), detemiine whether GC7MS confirmation was performed. If it was performed, qualify the data using the following guidance (see Table 15); a. If GC'/MS confirmation was not required because the quantitative criteria for both columns was not met, but it was still performed, use professional judgment when evaluating the data to decide w hether the detect should be qualified with "C '. b. If GC/MS confirmation was performed, but unsuccessful for a target compound detected by GC/TiCD analysis, qualify those detects as "X". Table 15. (,as Chronialo^aph/.Ylass Spectrometer ((.(VMS) Confirmation Actions Criteria Action SCP/Toxaphene was confirmed by GC/MS Detects C SCP/Toxaphene was not con tinned by GC/MS Detects X 32 ------- SOP IIW-36A Revision 0 June 2015 Compound Quantitation and Reported Contract Required Quantitation Limits (CROl.s) Action: 1. When a sample is analy/ed at more than one dilution, the lowest CRQLs are used unless a QC exceedance dictates the use of the higher CRQLs from the diluted sample. Replace concentrations that exceed the calibration range in the original analysis by crossing out the "ir and its corresponding value on the original Form 1 and substituting the data from the diluted sample. 2. Results between the MDL and CRQL should be qualiiled as estimated (J). 3. Results less than the MDL should be reported at the CRQL and qualified (U). MDLs themselves are not reported. 4. For non-aqueous samples, if the percent moisture is less than 70.0%. no qualification of the data is necessary. If the percent moisture is greater than or equal to 70.0% and less than 90.0° o, qualify detects as estimated (J) and non-detects as approximated (UJ). If the percent moisture is greater than or equal to 90.0°o, qualify delects as estimated (J) and non-detects as unusable (R) (see Table 16). 5. If any discrepancies are found, the Region's designated representative may contact the laboratory to obtain additional information that could resolve any differences. If a discrepancy remains unresolved, the reviewer must use professional judgment to decide which value is the most accurate. Under these circumstances, the reviewer may determine that qualification of data is warranted. Note in the Data Review Narrative a description of the reasons for data qualification and the qualification that is applied to the data. 6. Note, for Contract I.aboratory Program Project Officer (CLP PO) action, numerous or significant failures to accurately quantify the target compounds or to properly evaluate and adjust CRQLs. Table 16. Percent Moisture Actions for Pesticide Anahsis For Non-Aqueous Samples Action Criteria Detected Associated Non-detected Associated Compounds Compounds % Moisture 70.0 No qualification 70.0 ^ °o Moisture 90.0 J UJ % Moisture > 90.0 J R 33 ------- SOP IIW-36A Revision 0 June 2015 Field Duplicates Action; NOTE: In the absence of QAPP guidance lor validating data Irom field duplicates, the following action will be taken. Identity which samples within the data package are Held duplicates. Hstimate the relative percent difference (RPD) between the values for each compound. If large RPDs (> 50%) is observed, confirm identification of samples and note difference in the executive summary. 34 ------- SOP IIW-36A Revision 0 June 2015 Overall Assessment oI'Daia Action; 1. Use professional judgment to determine if there is any need to quali I'v data which were not qualified based on the Quality Control (QC) criteria previously discussed, 2. Write a brief narrative to give the user an indication ofthe analytical limitations ofthe data. Note, for Contract Laboratory Program Project Officer (CLP PO) action, any inconsistency ofthe data with the Sample Delivery Group (SDG) Narrative. If sufficient information on the intended use and required quality ofthe data is available, the reviewer should include their assessment ofthe usability ofthe data within the given context. This may be used as part of a formal Data Quality Assessment (DQA). 35 ------- SOP IIW-36A Revision 0 June 2015 APPENDIX A: GLOSSARY Analytc — The element of interest, ion, or parameter an analysis seeks to determine. Analytical Services Branch (ASH) — Directs the Contract Laboratory Program (CLP) from within the Office of Superfund Remediation and Technical Innovation (OSRTI) in the Office of Solid Waste and Emergency Response (OSWHR). Analytical Sample — Any solution or media introduced into an instrument on which an analysis is performed excluding instrument calibration, Initial Calibration Verification (ICV), Initial Calibration Blank (ICR). Continuing Calibration Verification (CCV), and Continuing Calibration Blank (CCB). Note that the following are all defined as analytical samples: undiluted and diluted samples (USHPA and non-US FP A); Matrix Spike samples; duplicate samples; serial dilution samples, analytical (post-digestion/post-distillation) spike samples; Interference Check Samples (ICSs); Laboratory Control Samples (LCSs); and Preparation Blanks. Associated Samples -- Any sample related to a particular Quality Control (QC) analysis. For example, for Initial Calibration Verification (ICV), all samples run under the same calibration curve. For duplicates, all Sample Delivery Group (SDG) samples digested/distilled of the same matrix. Blank — A sample designed to assess specific sources of contamination. See individual definitions for types of blanks. Calibration -- The establishment of an analytical curve based on the absorbance, emission intensity, or other measured characteristic of known standards. The calibration standards are to be prepared using the same type of reagents or concentration of acids as used in the sample preparation. Calibrat ion Blank — A blank solution containing all of the reagents in the same concentration as those used in the analytical sample preparation. This blank is not subject to the preparation method. Calibration Curve -- A plot of instrument response versus concentration of standards. Calibration Standards -- A series of known standard solutions used by the analyst for calibration of the instrument (i.e., preparation of the analytical curve). The solutions may or may not be subjected to the preparation method, but contain the same matrix (i.e.. the same amount of reagents and/or preservatives) as the sample preparations to be analyzed. Case -- A Unite, usually predetermined number of samples collected over a given lime period from a particular site. Case numbers are assigned by the Sample Management Office (SMO). A Case consists of one or more Sample Delivery Groups (SDGs). Contract Compliance Screening (CCS) — A screening of electronic and hardcopy data deliverables for completeness and compliance with the contract. This screening is performed under USLPA direction by the Contract Laboratory Program (CLP) Sample Management Office (SMO) contractor. Continuing Calibration Verification (CCV) -- A single parameter or multi-parameter standard solution prepared by the analyst and used to verify the stability ofthe instrument calibration with time, and the instrument performance during the analysis of samples. The CCV can be one ofthe calibration standards. I lowever, all parameters being measured by the particular system must be represented in this standard and the standard must have the same matrix (i.e.. the same amount of reagents and/or preservatives) as the samples. 36 ------- SOP IIW-36A Revision 0 June 2015 Contract Laboratory Program (CLP) ~ Supports the USLPA's Super fund effort by providing a range of state-of-the-art chemical analytical services of known quality. This program is directed by the Analytical Services Branch (ASB) of the Office of Superlund Remediation and Technical Innovation (OSRTI) of USf.PA. Contract Laboratory Program Project Officer (CLP PO) — 1 ho Regional I SI PA official responsible for monitoring laboratory performance and/or requesting analytical data or services from a CLP laboratory. Contract Required Quantitation Limit (CRQL) — Minimum level of quantitation acceptable under the contract Statement of Work (SOW). Duplicate - A second aliquot of a sample that is treated the same as the original sample in order to determine the precision oflhe method. Field Blank — Any sample that is submitted from the Held and identified as a blank. A Held blank is used to check for cross-contamination during sample collection, sample shipment, and in the laboratory. A field blank includes trip blanks, rinsate blanks, bottle blanks, equipment blanks, preservative blanks, decontamination blanks, etc. Field Duplicate — A duplicate sample generated in the Held, not in the laboratory. Holding Time — The maximum amount of time samples may be held before they are processed. Contractual - The maximum amount oftime that the Contract I.aboratory Program (CLP) laboratory may hold the samples from the sample receipt dale until analysis and still be in compliance with the terms of the contract, as specified in the CLP Analytical Services Statement of Work (SOW), t hese times are the same or less than technical holding times to allow for sample packaging and shipping. Technical — The maximum amount of time that samples may be held from the collection date until analysis. Initial Calibration -- Analysis of analytical standards for a series of different spec i lied concentrations to define the quantitative response, linearity, and dynamic range of the instrument to target analvtes. Initial Calibration Verification (ICV) — Solution(s) prepared from stock standard solutions, metals, or salts obtained from a source separate from that utilized to prepare the calibration standards. The ICV is used to verify the concentration of the calibration standards and the adequacy of the instrument calibration. The ICV should be traceable to National Institute of Standards and Technology (NIST) or other certified standard sources when USKPA ICV solutions are not available. Internal Standard — A non-target element added to a sample at a known concentration after preparation but prior to analysis. Instrument responses to internal standards are monitored as a means ofassessing overall instrument performance. Matrix — The predominant material of which the sample to be analyzed is composed, for the purposes of this document, the matrices are aqueous/water, soil/sediment, wipe, and filter. Matrix Spike — Introduction of a known concentration of analyte into a sample to provide information about the effect ofthe sample matrix on the digestion and measurement methodology (also identified as a pre-distillation/digestion spike). Method Detection Limit (.MDL) ~ The concentration of a target parameter that, when a sample is processed through the complete method, produces a signal with 99 percent probability that it is different from the blank, for 7 replicates ofthe sample, the mean value must be 3.14s above the blank, where "s" is the standard deviation ofthe 7 replicates. 37 ------- SOP IIW-36A Revision 0 June 2015 Narrative (SI)G Narrative) -- Portion of the data package which includes laboratory, contract, Case, Sample Number identification, and descriptive documentation of any problems encountered in processing the samples, along with corrective action taken and problem resolution. Office of Solid Waste and Emergency Response (OSWER) - The I SI l'A office that provides policy, guidance, and direction for the E'SEPA's solid waste and emergency response programs, including Siiperfund. Percent Difference (%D) — As used in this document and the Statement of Work (SOW), is used to compare two values. The difference between the tw o values divided by one of the values. Performance Evaluation (PE) Sample — A sample of known composition provided by I ;SEPA for contractor analysis. Used by USE PA to evaluate C ontractor performance. Preparation Blank — An analytical control that contains reagent water and reagents, which is carried through the entire preparation and analytical procedure. Relative Percent Difference (RPD) — As used in this document and the Statement of Work (SOW) to compare two values, the RPD is based on the mean of the two values, and is reported as an absolute value (i.e., always expressed as a positive number or zero). Regional Sample Control Center Coordinator (RSCC) -- In I Si PA Regions, coordinates sampling efforts and serves as the central point-of-contact for sampling questions and problems. Also assists in coordinating the level of Regional sampling activities to correspond with the monthly projected demand for analytical services. Relative Standard Deviation (RSD) — As used in this document and the Statement of Work (SOW), the mean divided by the standard deviation, expressed as a percentage. Sample — A single, discrete portion of material to be analyzed, which is contained in single or multiple containers and identified by a unique Sample Number. Sample Delivery Croup (SDC) - A unit within a sample Case that is used to identify a group of samples for delivery. An SI)G is defined by the following, whichever is most frequent: a. Each 20 field samples [excluding Performance F;valuation (PH) samples] within a Case; or b. Each 7 calendar day period (3 calendar day period for 7-day turnaround) during which field samples in a Case are received (said period beginning with the receipt of the first sample in the SIXi). c. Scheduled at the same level of deliverable. In addition, all samples and/or sample fractions assigned to an SIXi must be scheduled under the same contractual turnaround time. Preliminary Results have no impact on defining the SIX3. Samples may be assigned to SDGs by matrix (i.e., all soil/sediment samples in one SIXi. all aqueous/water samples in another) at the discretion of the laboratory. Sample Management Office (SMO) -- A contractor-operated facility operated under the SMO contract, awarded and administered by the USEPA. Provides necessary management, operations, and administrative support to the Contract Eaboratory Program (CEP). Statement of Work (SOW) — A document which specifies how laboratories analyze samples under a particular Contract Eaboratory Program (CEP) analytical program. 38 ------- SOP HW-36A Revision 0 June 2015 APPENDIX B: ORGANIC DATA EXECUTIVE NARRATIVE TEMPLATE UNITED STATES ENVIRONMENTAL PROTECTION AGENCY REGION 2 DESA/HWSBJHWSS 2890, Wood bridge Avenue, Edison, NJ 08837 EXECUTIVE NARRATIVE Case No.: S ite:| Number of Samples: Analysis: SDG No.: Laboratory: Sampling dates: QAPP HWSS ft Contractor Document #: SUMMARY: Critical: Results have an unacceptable level of uncertainty and should riot be used for making decisions Data have been qualified "R" rejected. Major: A level of uncertainty exists that may not meet the data quality objectives for the project A bias is likely to be present in the results. Data has been qualified "J' estimated. Minor: The level of uncertainty is acceptable. No significant bias in the data was observed. Critical Findings: Major Findings: Minor Findings: Reviewer Name(s): Approver's Signature: Name: Affiliation: Date: USEPA/R2i'HWSBiHWSS Page 1 of 3 39 ------- SOP IIW-36A Revision 0 June 2015 AIM'KNDIX C : SAM PLC ORCANIC DATA SAM I'll SUMMARY f.1-r-NV OQOCi Contract: SDGNo- • - laHCod? ocooi sample Nuj.tbei BA!X'< M.-tho-J V>iln\ veil MA Number: Sainpl-1 ocation pll S 2 Saittpb Drtr i- * Sain] k» Time 13 'JJ "o Mos^mre :£• r> ~i Willi* Analyte Name Result Units Dilution Factor Lab Flag Validation Reportable Validation Level 2.4.5-T ir, ug kg 1,0 i; r Ycv S5YEM J " ^-T? i Mb. c\< -Y" ug kg 1,0 IT U Yes S3YEM - —t- 59 15 L" k 2 i.O u u \ C> S3YEM ; 4-rm 39 »k, kj 1.0 u r Ye> ViYF.M Dal1!'' it. 97 WL 1 J u u Y« DiCarib1 •' re kg i Yes •vVf-M liy kj [ • \ e* \ATN! Dm^eb 1L ] nuLj u u Yes VYfcM Mi. PA :*-> ) kj u UJ Yc- S.iYEM Mt PP ug'ke r o Yes SJYEM Pwiit whk'ij|ii ai ol 3,9 lit. U iT u Yc- S.;VEM 4-\ns^j'i cnoi 39 HC 1'4 1,0 u u Yes ViVfM 40 ------- SOP IIW-36A Revision 0 June 2015 APPENDIX I): 1LEC TRONIC DATA DELIVERABLE TEMPLATE DATA PROVIDER 1 AB MA I RIX CODI RIST LI UNI I SYS SAMPLH CODE ANAL I OCA DON DETECTION LIMIT UNIT SAMPLE NAME BASIS ITC RETENTION TIME SWUM ! MA I RIX CODE CONTAINER ID RI.SUI 1 COMMLN'I SAM PI E. TV PI- CODE Din iton LAcroR QC ORICilNAI. CONC SAMPI I- SOl.RCT. PRI'P METHOD QC SPIKE ADDED PAREN r SAMPI I-: CODE PREP DATE OC SPIKE' MEASURED SAMPLE 1)1'I. (iROl.'P LLACflATL METHOD OC SPIKE RECOVERY SAMPI i . DA IT. 1J-ACHATE date: QC DUP ORKilN \I. CONC SYS l,OC CODI 1 AB NAME. CODI OC DI P SPIKE ADDET) STAR!' DEPTH QC LEVEL QC DUP SP1KI Ml \SI RI D end i)i-;i'ni LAB SAMPLE ID OC Dl P SPIKI REX'OVE'RY DEPIII CM 1 PERCENI MOISTURE OC RPD CHAIN ()l- Cl'SIODV SLBSAMPi.E AMOl NT OC SPIKE I CL SI N 1 iO I AM PA it: SI BSAMPLI AMOENI UNIT OC SPIKE UCL SAMPI I RECEIPT DA IK ANALYST NAME: OC RPD CL SAMPI. I. R INSTRUMENT ID OC SPIKE: STATUS SAMPI.INC.! COMPANY CODE COMMENT OC DUP SPIKE S I A 1 I S SAMPI INC) REASON preservative: OC RPD SI A IT'S SAMIM INC I K 11NlOt ! TINA! YOLl ME BREAK 2 I ASK com-: TINA I VOl I Ml. UNI 1 sys sample: code: COI I.I CI ION Ql.'ARITR CAS RN I AB AM. ME 1 HOD NAME COMPOSIIE YN CI 1EMICAI NAME A N AI YSIS DA IT COMPOSI I 1 DI SC RISLEI valt e: TOTAL OR DISSOLVED SAMPLE CLASS re:sult error delta COLUMN NUMBER CUSTOM FIELD 1 result type: code: TEST TYPE CUSTOM FIELD 2 REPORTABLE RESULT TEST BATCH TYPE CI SIOM ITI I D DEI LCI T1 .Ati TEST BATCH ID COM Ml-NT LAB 01 A.I ITII RS CASE BREAK 1 VA1 IDA TOR 01 A1 IT 1! RS CONTRACT NUM SYS SAMPLE CODE INTERPRE 1 ED QUAI 1M1 RS SCRIBE SAMPLE. ID 1 AB ANI. ML I I101) NAME ORdANIC YN sample: time ANALYSIS DA IT METHOD DEI! (,'TION 1 IMI I T'RAC ITON TOTAL OR DISSOLVED REPOR 1 INti DETECTION LIMIT Pll con mn number QUANTITATION 1 IMI T DATA YAL LABEL TEST TYPE 41 ------- |