CRA	UNITED STATES ENVIRONMENTAL PROTECTION AGENCY

CrM	NATIONAL EXPOSURE RESEARCH LABORATORY

Environmental Protection	EXPOSURE METHODS AND MEASUREMENT DIVISION (MD-D205-03)

Agency	Research Triangle Park, NC 27711	Office of

Research and Development

LIST OF DESIGNATED REFERENCE AND EQUIVALENT METHODS

Issue Date: December 17, 2016

(www.epa.gov/ttn/amtic/criteria.html)

These methods for measuring ambient concentrations of specified air pollutants have been designated as
"reference methods" or "equivalent methods" in accordance with Title 40, Part 53 of the Code of Federal
Regulations (40 CFR Part 53). Subject to any limitations (e.g., operating range or temperature range) specified in
the applicable designation, each method is acceptable for use in state or local air quality surveillance systems under
40 CFR Part 58 unless the applicable designation is subsequently canceled. Automated methods for pollutants other
than PMio are acceptable for use only at shelter temperatures between 20°C and 30°C and line voltages between
105 and 125 volts unless wider limits are specified in the method description.

Prospective users of the methods listed should note (1) that each method must be used in strict accordance
with its associated operation or instruction manual and with applicable quality assurance procedures, and (2) that
modification of a method by its vendor or user may cause the pertinent designation to be inapplicable to the method
as modified. (See Section 2.8 of Appendix C, 40 CFR Part 58 for approval of modifications to any of these methods
by users.)

Further information concerning particular designations may be found in the Federal Register notice cited
for each method or by writing to the National Exposure Research Laboratory, Exposure Methods and Measurement
Division (MD-D205-03), U.S. Environmental Protection Agency, Research Triangle Park, North Carolina 27711.
Technical information concerning the methods should be obtained by contacting the source listed for each method.
Source addresses are listed at the end of the listing of methods, except for the addresses for lead method sources,
which are given with the method. New analyzers or PMio samplers sold as reference or equivalent methods must
carry a label or sticker identifying them as designated methods. For analyzers or PMio samplers sold prior to the
designation of a method with the same or similar model number, the model number does not necessarily identify
an analyzer or sampler as a designated method. Consult the manufacturer or seller to determine if a previously sold
analyzer or sampler can be considered a designated method or if it can be upgraded to designation status. Analyzer
users who experience operational or other difficulties with a designated analyzer or sampler and are unable to
resolve the problem directly with the instrument manufacturer may contact EPA (preferably in writing) at the above
address for assistance.

This list will be revised as necessary to reflect any new designations or any cancellation of a designation
currently in effect. The most current revision of the list will be available for inspection at EPA's Regional Offices,
and copies may be obtained at the Internet site identified above or by writing to the National Exposure Research
Laboratory at the address specified above.


-------
Most Recent Designations



Sutron Model 6020 Sulfur Dioxide Fluorescent Analyzer

July 2016

Teledyne API Model T640 PM mass monitor with 640X option (PM10, PM2 sand PM10-25)

July 2016

Teledyne API Model T640 PM mass monitor (PM2 5)

July 2016

Teledyne API Model T200P Chemiluminescence Nitrogen Oxides Analyzer

October 2016


-------
TSP

List of Designated Reference and Equivalent Methods. December 17.2016

Page 3

Particulate Matter - TSP

Reference Method for TSP

Manual Reference Method: 40 CFR Part 50, Appendix B

Reference Method for the Determination of Suspended Particulate Matter in the Atmosphere (High-Volume Method)
Federal Register: Vol. 47, page 54912,12/06/82 and Vol 48, page 17355, 04/22/1983


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 4

Particulate Matter - PM10

Andersen Model RAAS10-100 PMto Single Channel PMio Sampler
Manual Reference Method: RFPS-0699-130

"Andersen Instruments, Incorporated Model RAAS10-100 Single Channel Reference Method PMio Sampler," with RAAS-10 PMio
inlet or the louvered inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a PMio reference method, and
operated for a continuous 24-hour sample period at a flow rate of 16.67 Lpm, and in accordance with the Model RAAS105-100
Operator's Manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix J.

Federal Register: Vol. 64, page 33481, 06/23/1999

Andersen Model RAAS10-200 PMio Single Channel PMio Audit Sampler
Manual Reference Method: RFPS-0699-131

"Andersen Instruments, Incorporated Model RAAS10-200 Single Channel Reference Method PMio Audit Sampler," with
RAAS-10 PMio inlet or the louvered inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a PMio
reference method, and operated for a continuous 24-hour sample period at a flow rate of 16.67 Lpm, and in accordance with
the Model RAAS105-200 Operator's Manual and with the requirements and sample collection filters specified in 40 CFR Part
50, Appendix J.

Federal Register: Vol. 64, page 33481, 06/23/1999

Andersen Model RAAS10-300 PMioMulti Channel PMio Sampler
Manual Reference Method: RFPS-0699-132

"Andersen Instruments, Incorporated Model RAAS10-300 Multi Channel Sequential Reference Method PMio Sampler," with
RAAS-10 PMio inlet or the lowered inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a PMio
reference method, and operated for a continuous 24-hour sample period at a flow rate of 16.67 Lpm, and in accordance with
the Model RAAS105-300 Operator's Manual and with the requirements and sample collection filters specified in 40 CFR Part
50, Appendix J.

Federal Register: Vol. 64, page 33481, 06/23/1999

BGI Incorporated Model PQ100 Air Sampler
Manual Reference Method: RFPS-1298-124

"BGI Incorporated or Mesa Laboratories Incorporated Model PQ100 Air Sampler," with BGI 16.7 Inlet Kit or the lowered
inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a PMio reference method, for a continuous 24-
hour sample period at a flow rate of 16.7 Lpm, with original firmware Version 5.X and lower or new firmware version 6.0 and
higher, operated in accordance with the original Model PQ100 Instruction Manual or manual revision Version 7.0, as
appropriate, and with the requirements specified in 40 CFR Part 50, Appendix J, using either the original or the newer PQ200-
type filter cassettes, and with or without the optional Solar Panel Power Supply.

Federal Register: Vol. 63, page 69625,12/17/1998
Latest modification: 01/2009; 6/2015

BGI Incorporated Model PQ200 Air Sampler
Manual Reference Method: RFPS-1298-125

"BGI Incorporated or Mesa Laboratories Incorporated Model PQ200 Air Sampler," with "flat plate" PMio inlet or the lowered
inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a PMio reference method, and operated for a
continuous 24-hour sample period in accordance with the Model PQ200 Instruction Manual and with the requirements specified
in 40 CFR Part 50, Appendix J, and with or without the optional Solar Panel Power Supply.

Federal Register: Vol. 63, page 69625,12/17/1998
Latest modification: 6/2015

DKK-TOA Models FPM-222/222C, FPM223/223C, and DUB-222(S)/223(S) PMio Monitor
Automated Equivalent Method: EQPM-0905-156

"DKK-TOA Models FPM-222, FPM-222C, FPM-223, FPM-223C, DUB-222(S), and DUB-223(S) Particulate Monitor," for
monitoring PMio in Ambient Air (beta attenuation monitor), configured for PMio, with Firmware Version DUB4-658355,
Corrected Slope Factor (FACT SLOPE) set to 1.232, Corrected Zero Value (FACT ZERO) set to 1.8, and with or without any
of the following options: Auto Check and Serial Recorder.

Federal Register: Vol. 70, page 56684, 09/28/2005


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 5

Ecotech Model 3000 PMio High Volume Air Sampler
Manual Reference Method: RFPS-0706-162

"Ecotech Pty. Ltd. Model 3000 PMio High Volume Air Sampler," configured with the Ecotech PMio Size-Selective Inlet
(SSI)(P-ECO-HVS3000-02), with the flow rate set to 1.13 m3/min (67.8 m3/hour).

Federal Register: Vol. 71, page 42089, 07/25/2006

Environnement S.A. Model MP101M PMio Monitor
Automated Equivalent Method: EQPM-0404-151

"Environnement S. A. Model MP101M PMio Beta Gauge Monitor," configured with the louvered PMio inlet specified in 40
CFR 50 Appendix L or its flat-topped predecessor version and one of the three optional temperature-regulated sampling tubes
(RST), and operated with a full scale measurement range of 0 - 0.500 mg/m3 (0 - 500 :g/m3), with the sample flow rate set to
1.00 m3/h and flow regulation set to yes, the "norms selection" set to m3 (actual volume), the "cycle" set to 24 hours, the
"period" set to none, and the "counting time" set to 200 seconds.2
Federal Register: Vol. 69, page 18569, 4/8/2004

Graseby Andersen/GMW Model 1200 High-Volume Air Sampler
Manual Reference Method: RFPS-1287-063

"Sierra-Andersen or General Metal Works Model 1200 PMio High-Volume Air Sampler System," consisting of a Sierra-
Andersen or General Metal Works Model 1200 PMio Size-Selective Inlet and any of the high-volume air samplers identified
as SAUV-10H, SAUV-11H, GMW-IP-10, GMW-IP-10-70, GMW-IP-10-801, or GMW-IP-10-8000, which include the
following components: Anodized aluminum high-volume shelter with either acrylonitrile butadiene styrene plastic filter holder
and motor/blower housing or stainless steel filter holder and phenolic plastic motor/blower housing; 0.6 hp motor/blower;
pressure transducer flow recorder; either an electronic mass flow controller or a volumetric flow controller; either a digital
timer/programmer, seven-day mechanical timer, six-day timer/programmer, or solid-state timer/programmer; elapsed time
indicator; and filter cartridge.

Federal Register: Vol. 52, page 45684,12/01/1987 and Vol 53, page 1062, 01/15/1988

Graseby Andersen/GMW Model 321-B High-Volume Air Sampler
Manual Reference Method: RFPS-1287-064

"Sierra-Andersen or General Metal Works Model 321-B PMio High-Volume Air Sampler System," consisting of a Sierra-
Andersen or General Metal Works Model 321-B PMio Size-Selective Inlet and any of the high-volume air samplers identified
as SAUV-10H, SAUV-11H, GMW-IP-10, GMW-IP-10-70, GMW-IP-10-801, or GMW-IP-10-8000, which include the
following components: Anodized aluminum high-volume shelter with either acrylonitrile butadiene styrene plastic filter holder
and motor/blower housing or stainless steel filter holder and phenolic plastic motor/blower housing; 0.6 hp motor/blower;
pressure transducer flow recorder; either an electronic mass flow controller or a volumetric flow controller; either a digital
timer/programmer, seven-day mechanical timer, six-day timer/programmer, or solid-state timer/programmer; elapsed time
indicator; and filter cartridge.

Federal Register: Vol. 52, page 45684,12/01/1987 and Vol 53, page 1062, 01/15/1988

Graseby Andersen/GMW Model 321-C High-Volume Air Sampler
Manual Reference Method: RFPS-1287-065

"Sierra-Andersen or General Metal Works Model 321-C PMio High-Volume Air Sampler System," consisting of a Sierra-
Andersen General Metal Works Model 321-C PMio or Size-Selective Inlet and any of the high-volume air samplers identified
as SAUV-10H, SAUV-11H, GMW-IP-10, GMW-IP-10-70, GMW-IP-10-801, or GMW-IP-10-8000, which include the
following components: Anodized aluminum high-volume shelter with either acrylonitrile butadiene styrene plastic filter holder
and motor/blower housing or stainless steel filter holder and phenolic plastic motor/blower housing; 0.6 hp motor/blower;
pressure transducer flow recorder; either an electronic mass flow controller or a volumetric flow controller; either a digital
timer/programmer, seven-day mechanical timer, six-day timer/programmer, or solid-state timer/programmer; elapsed time
indicator; and filter cartridge.

Federal Register: Vol. 52, page 45684,12/01/1987 and Vol 53, page 1062, 01/15/1988


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 6

Graseby Andersen/GMWModels SA241 and SA241MDichotomous Sampler
Manual Reference Method: RFPS-0789-073

"Sierra-Andersen Models SA241 and SA241M or General Metal Works Models G241 and G241M PMio Dichotomous
Samplers," consisting of the following components: Sampling Module with SA246b or G246b 10 nm inlet or the louvered inlet
specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, 2.5 nm virtual impactor assembly, 37 mm coarse and fine particulate
filter holders, and tripod mount; Control Module with diaphragm vacuum pump, pneumatic constant flow controller, total and
coarse flow rotameters and vacuum gauges, pressure switch (optional), 24-hour flow/event recorder, digital timer/programmer
or 7-day skip timer, and elapsed time indicator.

Federal Register: Vol. 54, page 31247, 07/27/1989

Graseby Andersen/GMW Model FH621-N Beta Monitor
Automated Equivalent Method: EQPM-0990-076

"Andersen Instruments Model FH62I-N PMio Beta Attenuation Monitor," consisting of the following components: FH101
Vacuum Pump Assembly; FH102 Accessory Kit; FH107 Roof Flange Kit; FH125 Zero and Span PMio Mass Foil Calibration
Kit; FH62I Beta Attenuation 19-inch Control Module; SA246b PMio Inlet (16.7 Lpm) or the louvered inlet specified in 40 CFR
50 Appendix L, Figs. L-2 thru L-19; operated for 24-hour average measurements, with an observing time of 60 minutes, the
calibration factor set to 2400, a glass fiber filter tape, an automatic filter advance after each 24-hour sample period, and with or
without either of the following options: FH0P1 Indoor Cabinet; FH0P2 Outdoor Shelter Assembly.

Federal Register: Vol. 55, page 38387, 09/18/1990

Met One or Sibata Models BAM/GBAM1020, BAM/GBAM1020-1, Horiba APDA-3 71, or Ecotech Spirant BAM1000
Automated Equivalent Method: EQPM-0798-122

"Met One Instruments or Sibata Scientific Technology Models BAM 1020, GBAM 1020, BAM 1020-1, GBAM 1020-1, Horiba
APDA-371 PMio Beta Attenuation Monitor, or Ecotech Spirant BAM1000" including the BX-802 EPA PMio inlet (or
alternative lowered PMio inlet meeting 40 CFR 50 Appendix L specifications), operated for 24-hour average measurements,
with a filter change frequency of one hour, with glass fiber filter tape, and with or without any of the following options: BX-
823, tube extension; BX-825, heater kit; BX-826, 230 VAC heater kit; BX-827 "Smart Heater" set for maintaining moisture
between 35% and 45% and no AT control; BX-828, roof tripod; BX-902, exterior enclosure; BX-903, exterior enclosure with
temperature control; BX-961, mass flow controller; BX-967, internal calibration device, BX-970 touch-screen display with
USB interface. For software (firmware) versions V3.0 or higher, a user-selectable measurement time (COUNT TIME) of 4, 6,
8 or 10 minutes selected, along with appropriate sample time (BAM SAMPLE) setting of 50,46,42 or 38 minutes, respectively,
to maintain a 60-minute measurement cycle. For software (firmware) versions V3.5 or higher, user-selectable option to sample
under actual conditions (Flow Type: ACTUAL) and report under standard conditions (Reporting: STD), which requires the use
of P/N BX-592 external temperature sensor or P/N BX-596 external temperature/barometric pressure sensor. The user may
also sample under standard conditions (Flow Type: STD) and report under standard conditions (Reporting: STD) with any
software/firmware 2.0 or higher. Instrument must be operated in accordance with the appropriate instrument manual.

Federal Register: Vol. 63, page 41253, 08/03/1998
Latest modifications: 06/2009; 07/2010; 8/2010; 8/2012

Met One Instruments, Inc. E-BAMPLUS Beta Attenuation Mass Monitor- PMio FEM Configuration
Automated Equivalent Method: EQPM-1215-226

"Met One Instruments, Inc. E-BAM PLUS Beta Attenuation Mass Monitor - PMio FEM Configuration," configured for 24 1-
hour average measurements of PMio by beta attenuation, using a glass fiber filter tape roll (460130 or 460180), a sample flow
rate of 16.67 liters/min, with the standard (BX-802) EPA PMio inlet (meeting 40 CFR 50 Appendix L specifications) and
equipped with 9250 ambient temperature sensor. Instrument must be operated in accordance with the E-BAM PLU S Particulate
Monitor operation manual, revision 1 or later. This designation applies to PMio measurements only.

Federal Register: Vol. 81, page 4294, 01/26/2016

Met One Instruments, Inc. E-FRM- PMio
Manual Reference Method: RFPS-0216-231

"Met One Instruments, Inc. E-FRM," configured for filter sampling of ambient particulate matter using the US EPA PMio
inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, with a flow rate of 16.67 L/min, using 47 mm PTFE membrane
filter media, and operating with firmware version R2.0.1 and later, and operated in accordance with the Met One E-FRM PM10
operating manual. This designation applies to PMio measurements only.

Federal Register: Vol. 81, page 25397, 04/28/2016


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 7

Opsis Model SM200 PMio Monitor
Automated Equivalent Method: EQPM-0810-193

"Opsis Model SM200 Monitor," beta gauge semi-continuous ambient particulate monitor operated for 24 hours at a flow rate
of 16.67 Lpm between 5° and 40°C using 47 mm PTFE membrane filter media, in the mass measurement range of 0 to 60 mg,
configured with a BGI Model SSI25 PMio inlet meeting criteria specified in 40 CFR 50 Appendix L, with a roof mounting kit,
and with or without an inlet tube heater (as recommended based on site RH conditions), according to the SM200 User's Guide.
Federal Register: Vol. 75, page 51039, 08/18/2010

Oregon DEQ Medium Volume PMio Sampler
Manual Reference Method: RFPS-0389-071

"Oregon DEQ Medium Volume PMio Sampler." NOTE: This method is no longer commercially available.

Federal Register: Vol. 54, page 12273, 03/24/1989

Teledyne Model 602 BetaPLUS Particle Measurement System or SWAM 5a Dual Channel Monitor
Automated Equivalent Method: EQPM-0912-205

"Teledyne Model 602 Bcta'"/ , ;s Particle Measurement System" or "SWAM 5a Dual Channel Monitor" configured for 1-hour
measurements of PMio by beta attenuation on a single sampling line (Line A or B, but not both together), with the standard,
louvered US EPA PMio size selective inlet specified in 40 CFR 50 Appendix L, using 47 mm glass fiber filters, at a sample
flow set to 16.67 liters/min and software version 05-02.07.63 or later. Operated in accordance with the Teledyne Model 602
Bcta'"/ , :s Particle Measurement System Operation Manual.

Federal Register: Vol. 77, page 60985,10/5/2012

Teledyne Advanced Pollution Instrumentation Model T640 PM mass monitor with 640X option
Automatic Equivalent Method: EQPM-0516-239

"Teledyne Advanced Pollution Instrumentation Model T640 PM mass monitor with 640X option," continuous ambient
particulate monitor operated at a volumetric flow rate of 16.67 Lpm, equipped with the louvered PMio inlet specified in 40
CFR 50 Appendix L, Figs. L-2 thru L-19, TAPI aerosol sample conditioner (P/N: 081040000), configured for operation with
firmware version 1.0.2.126 or later, in accordance with the Teledyne Model T640 Operations Manual. This designation
applies to PMio measurements only.

Federal Register: Vol. 81, page 45285, 07/13/2016

Thermo Andersen Series FH 62 C14 Continuous PM in Monitor

Thermo Scientific Model 5014i Beta (5014i Beta), Continuous Ambient Particulate Monitor
Automated Equivalent Method: EQPM-1102-150

"Thermo Andersen Series FH 62 C14 Continuous PMio Ambient Particulate Monitor and Thermo Scientific Model 5014i Beta
(5014i Beta), Continuous Ambient Particulate Monitor," operated for 24-hour average measurements, with the specified 10-
micron EPA PMio inlet (or alternative lowered PMio inlet meeting 40 CFR 50 Appendix L specifications), inlet connector,
sample tube with heater, roof flange kit, mass foil kit, pump kit, sample filter tape; with operational settings of 1000 L/h (16.67
L/min) sample flow rate, daily filter change, auto filter change at volumetric flow <950 L/h, auto filter change at mass >1500
micrograms, and factory default calculation mode settings operated with software version 1.07. Operated, calibrated and
serviced according to the appropriate Operator Manual.

Federal Register: Vol. 67, page 76174,12/11/2002
Latest modifications: 07/2009; 12/2012

Thermo Scientific or Rupprecht & Patashnick Partisol" Model 2000 Air Sampler
Manual Reference Method: RFPS-0694-098

"Thermo Scientific Partisol® 2000 Air Sampler" or "Rupprecht & Patashnick Partisol® Model 2000 Air Sampler," consisting
of a Hub Unit and 0, 1, 2, or 3 Satellite Units, with each sampling station used for PMio measurements equipped with a
Rupprecht & Patashnick PMio inlet and operated for continuous 24-hour sampling period using the Basic, Manual, Time,
Analog Input, or Serial Input programming modes, and with or without any of the following options: PM2 5-style filter cassette
holder; lowered inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19 in lieu of standard inlet; 57-002320 Stand for
Hub or Satellite; 59-002542 Advanced EPROM; 10-001403 Large Pump (1/4 hp); 120 VAC. Hardware for Indoor Installation
consists of: 51-002638-xxxx Temperature Sensor (Extended Length); 55-001289 Roof Flange (1 1/4"); 57-000604 Support
Tripod for Inlet; 57-002526-0001 Sample Tube Extension (1 m); 57-002526-0002 Sample Tube Extension (2 m). Hardware for
Outdoor Installation in Extreme Cold Environments consists of: 10-002645 Insulating Jacket for Hub Unit.

Federal Register: Vol. 59, page 35338, 07/11/1994


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 8

Thermo Scientific Partisof 2000-D Dichotomous Air Sampler or Thermo Fisher Scientific Partisof 2000i-D Dichotomous Air Sampler
Manual Equivalent Method: EQPS-0311-197

"Thermo Scientific Partisol® 2000-D Dichotomous Air Sampled' or "Thermo Fisher Scientific Partisol® 2000/-D Dichotomous Air Sampler,"
configured for dual-filter, single-event sampling of fine (PM25) and coarse (PMms) particles, operated with a U.S. EPA PM10 inlet and using a
virtual impactor to separate fine and coarse PM into two samples for collection on two separate filter membranes, for a 24-hour sampling period
and in accordance with the Partisol® 2000-D or Partisol® 2000/-D instruction manual, as appropriate. Partisol® 2000/-D operated with firmware
version 2.0 or greater.

Federal Register: Vol 76, page 15974,03/22/2011
Latest modification: 06/2011

Thermo Scientific Partisol® 2000-FRM PM10 Air Sampler or Thermo Fisher Scientific Partisol® 2000i PM10 Air Sampler
or Rupprecht andPatashnick Partisol-FRM 2000 PMioAir Sampler
Manual Reference Method: RFPS-1298-126

"Thermo Scientific Partisol® 2000-FRM PM10 Air Sampler" or "Thermo Fisher Scientific Partisol® 2000/ PM10 Air Sampler"
or "Rupprecht and Patashnick Partisol®-FRM 2000 PM10 Air Sampler," with PM10 inlet or louvered inlet specified in 40 CFR
50, Appendix L, Figs. L-2 through L-19, configured as a PM10 reference method with a U.S. EPA PM10 inlet with straight
downtube adapter and operated for continuous 24-hour sample periods in accordance with the Partisol® 2000-FRM or Partisol®
2000/ instruction manual, as appropriate, and with the requirements specified in 40 CFR Part 50, Appendix J. Model 2000/
operated with firmware version 2.0 or greater.

Federal Register: Vol. 63, page 69625,12/17/1998
Latest modification: 06/2011

Thermo Scientific Partisol-Plus 2025 PM10 Sequential Air Sampler or Thermo Fisher Scientific Partisol 2025i PM10
Sequential Air Sampler or Rupprecht and Patashnick Partisol91-Plus 2025 PM10 Sequential Air Sampler
Manual Reference Method: RFPS-1298-127

"Thermo Scientific Partisol®-Plus 2025 Sequential Air Samplef' or "Thermo Fisher Scientific Partisol® 2025/ Sequential Air
Samplef' or "Rupprecht and Patashnick Partisol®-Plus 2025 PM10 Sequential Air Sampler," with PM10 inlet or louvered inlet
specified in 40 CFR 50, Appendix L, Figs. L-2 through L-19, configured as a PM10 reference method and operated for continuous
24-hour sample periods. Partisol®-Plus 2025 to be operated with any software version 1.003 through 1.5 and the Partisol® 2025/
with firmware version 2.0 or greater, with the modified filter shuttle mechanism, in accordance with the Partisol®-Plus 2025 or
Partisol® 2025/ instruction manual, as appropriate, and with the requirements specified in 40 CFR Part 50, Appendix J.

Federal Register: Vol. 63, page 69625,12/17/1998
Last modified: 06/2011

Thermo Scientific Dichotomous Partisol -Plus 2025-D Sequential Air Sampler or Thermo Fisher Scientific Dichotomous
Partisol8 2025i-D Sequential Air Sampler
Manual Equivalent Method: EQPS-0311-198

"Thermo Scientific Dichotomous Partisol®-Plus 2025-D Sequential Air Sampler" or "Thermo Fisher Scientific Dichotomous
Partisol® 2025/-D Sequential Air Sampler," configured for dual-filter sampling of fine (PM2 5) and coarse (PM10-25) particles, with
a U.S. EPA PM10 inlet and using a virtual impactor to separate the fine and coarse PM into two samples for collection on two
separate filter membranes, and operated with the modified filter shuttle mechanism implemented May 31, 2008, and firmware
version 1.500 or greater for the Partisol®-Plus 2025-D and version 2.0 or greater for the Partisol® 2025/-D, for continuous 24-hour
sample periods , in accordance with the Partisol®-Plus 2025-D or Partisol® 2025/-D instruction manual, as appropriate.

Federal Register: Vol. 76, page 15975, 03/22/2011
Latest modification: 06/2011

Thermo Scientific TEOM® 1400AB/TEOM® 1405 Ambient Particulate Monitor or
Rupprecht & Patashnick TEOM® Series 1400/1400a PMm Monitors
Automated Equivalent Method: EQPM-1090-079

"Thermo Scientific TEOM® 1400AB [PM10] Ambient Particulate Monitor" or "Rupprecht & Patashnick TEOM® Series 1400
and Series 1400a PM-10 Monitors," (including serial number prefixes 1400, 140A, 140AA, 140AB, 140AT, and 140UP,
1405A), consisting of the following components: TEOM® Sensor Unit; TEOM® Control Unit; Flow Splitter (3 Lpm sample
flow); Teflon-Coated Glass Fiber Filter Cartridges; Rupprecht & Patashnick PM-10 Inlet (part number 57-00596), Sierra-
Andersen Model 246b PM-10 Inlet (16.7 Lpm) or lowered inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19;
operated for 24-hour average measurements, with the total mass averaging time set at 300 seconds, the mass rate/mass
concentration averaging time set at 300 seconds, the gate time set at 2 seconds, and with or without any of the following options:


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 9

Tripod; Outdoor Enclosure; Automatic Cartridge Collection Unit (Series 1400a only); Flow Splitter Adapter (for 1 or 2 Lpm
sample flow). Thermo Scientific TEOM® 1405 Ambient Particulate Monitor with combined sensor and control units and
redesigned mass transducer and user interface, operated in accordance with the Thermo Scientific TEOM® 1405 instrument
manual.

Federal Register: Vol. 55, page 43406,10/29/1990
Latest modification: 12/2008


-------
PMio

List of Designated Reference and Equivalent Methods. December 17.2016

Page 10

Thermo Scientific TEOM® 1405-DFDichotomous Ambient Particulate Monitor with FDMS®

Automated Equivalent Method: EQPM-1013-208

"Thermo Scientific TEOM® 1405-DF Dichotomous Ambient Particulate Monitor with FDMS®," configured for dual filter
sampling of fine (PM2 5) and coarse particles using the US EPA PM10 inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru
L-19 and a virtual impactor, with a total flow rate of 16.67 L/min, fine sample flow of 3 L/min, and coarse sample flow rate of
1.67 L/min, and operating with firmware version 1.70 and later, operated with or without external enclosures, and operated in
accordance with the Thermo Scientific TEOM® 1405-DF Dichotomous Ambient Particulate Monitor Instruction Manual. This
designation applies to PM10 measurements only.

Federal Register: Vol. 78, page 67360,11/12/2013

Tisch Environmental Model TE-6070 PM10 High-Volume Air Sampler or New Star Environmental Model NS-6070 PM10

High-Volume Air Sampler

Manual Reference Method: RFPS-0202-141

"Tisch Environmental Model TE-6070 or New Star Environmental Model NS-6070 PM10 High-Volume Air Sampler,"
consisting of a TE-6001 PM10 size-selective inlet, 8" x 10" filter holder, aluminum outdoor shelter, mass flow controller or
volumetric flow controller with brush or brushless motor, 7-day mechanical off/on-elapsed timer or 11-day digital off/on-
elapsed timer, and any of the high volume sampler variants identified as TE-6070-BL or NS-6070-BL, TE-6070D or NS-
6070D, TE-6070D-BL orNS-6070-BL, TE-6070V or NS-6070V, TE-6070V-BL or NS-6070V-BL, TE-6070-DVorNS-6070-
DV, or TE-6070DV-BL or NS-6070DV-BL, with or without the optional stainless steel filter media holder/filter cartridge or
continuous flow/pressure recorder.

Federal Register: Vol. 67, page 15566, 04/02/2002

Tisch Environmental Model TE- Wilbur 10 Particulate Sampler - PM10
Manual Reference Method: RFPS-0714-216

"Tisch Environmental Model TE-WilburlO PM10 Low-Volume Air Particulate Sampler," consisting of a TE-PM10-D PM10
size-selective inlet, specified in 40 CFR 50 Appendix L Figs. L-2 thru L-19, configured as a PM10 reference method, operated
for continuous 24-hour sample period at a flow rate of 16.67 L/min, using 47 mm PTFE membrane filter media, operated with
or without the optional TE-W-600 Solar Panel Power Supply kit, operating with firmware version 1.70 or later, and operated
in accordance with the Tisch Environmental Model TE-WilburlO PM10 Low-Volume Air Particulate Sampler instruction
manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix J.

Federal Register: Vol. 79, page 49307, 08/20/2014
Latest modification: 6/2015

Wedding & Associates' or Thermo Environmental Instruments Inc. Model 600 PM10 High-Volume Sampler
Manual Reference Method: RFPS-1087-062

"Wedding & Associates' or Thermo Environmental Instruments, Inc. Model 600 PM10 Critical Flow High-Volume Sampler,"
consisting of the following W&A/TEII components: PM10 Inlet; Critical Flow Device; Anodized Aluminum Shelter; Blower
Motor Assembly for 115, 220 or 240 VAC and 50/60 Hz; Mechanical Timer; Elapsed Time Indicator; and Filter
Cartridge/Cassette, and with or without the following options: Digital Timer, 6 or 7 Day Timer, and 1 or 7 Day Pressure
Recorder.

Federal Register: Vol. 52, page 37366, 10/06/1987

Wedding & Associates' or Thermo Environmental Instruments Inc. Model 650 PMioBeta Gauge
Automated Equivalent Method: EQPM-0391-081

"Wedding & Associates' or Thermo Environmental Instruments, Inc. Model 650 PM10 Beta Gauge Automated Particle
Sampler," consisting of the following W&A/TEII components: Particle Sampling Module, PM10 Inlet (18.9 Lpm), Inlet Tube
and Support Ring, Vacuum Pump (115, 220 or 240 VAC and 50/60 Hz); and operated for 24-hour average measurements with
glass fiber filter tape.

Federal Register: Vol. 56, page 9216, 03/05/1991


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 11

Particulate Matter - PM2.5

Andersen Model RAAS2.5-200 PM2.5 Ambient Audit Air Sampler
Manual Reference Method: RFPS-0299-128

"Andersen Instalments, Incorporated Model RAAS2.5-200 PM2 5 Audit Sampler," configured as a PM2 5 reference method and
operated with software (firmware) version 4B, 5.0.1 - 6.09, 6.0A, or 6.0B, for a continuous 24-hour sample period at a flow
rate of 16.67 liters/minute, and in accordance with the Model RAAS2.5-200 Operator's Manual and with the requirements and
sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 64, page 12167, 03/11/1999

BGIInc. Models PQ200 or PQ200A PM2.5 Ambient Fine Particle Sampler
Manual Reference Method: RFPS-0498-116

"BGI Incorporated or Mesa Laboratories Incorporated Models PQ200 and PQ200A PM25 Ambient Fine Particle Sampler,"
operated with firmware version 3.88, 3.89R, 3.91, 3.91R and 5.62 for a continuous 24-hour sample period, in accordance with
the Model PQ200/PQ200A Instruction Manual and with the requirements and sample collection filters specified in 40 CFR
Part 50, Appendix L, and with or without the optional Solar Power Supply or the optional dual-filter cassette (P/N F-21/6) and
associated lower impactor housing (P/N B2027), where the upper filter is used for PM2 5. The Model PQ200A is described as
a portable audit sampler and includes a set of three carrying cases.

Federal Register: Vol. 63, page 18911, 04/16/1998
Latest modification: 6/2015; 06/2016; 12/2016

BGI Inc. Models PQ200-VSCC™ or PQ200A- VSCC™ PM25 Sampler

Manual Reference Method: RFPS-0498-116 or Manual Equivalent Method: EQPM-0202-142

"BGI Incorporated or Mesa Laboratories Incorporated Models PQ200-VSCC™ or PQ200A-VSCC™ PM25 Ambient Fine
Particle Sampler," configured with a BGI Very Sharp Cut Cyclone (VSCC™) particle size separator and operated with firmware
version 3.88, 3.89R, 3.91, 3.91R and 5.62 for a continuous 24-hour sample period, in accordance with the Model
PQ200/PQ200A Instruction Manual and VSCC™ supplemental manual and with the requirements and sample collection filters
specified in 40 CFR Part 50, Appendix L, and with or without the optional Solar Power Supply or the optional dual-filter
cassette (P/N F-21/6) and associated lower impactor housing (P/N B2027), where the upper filter is used for PM2 5. The Model
PQ200A VSCC™ is described as a portable audit sampler and includes a set of three carrying cases.

Federal Register: Vol. 67, page 15567, 04/02/2002
Latest modification: 6/2015; 06/2016; 12/2016

Environnement S.A. Model MP101M PM2.5 Monitor
Automated Equivalent Method: EQPM-1013-211

"Environnement S.A. Model MP101M PM2.5 Beta Attenuation Monitor" using a glass fiber filter tape roll, operated at a
sample flow rate of 16.67 liters/min for 24-hour average measurements of PM2 5, configured with the standard EPA PM10 inlet
(meeting 40 CFR 50 Appendix L specifications) associated with a BGI Very Sharp Cut Cyclone (VSCC™) particle size
separator and using a temperature regulated sampling tube with ambient meteorological sensor. With or without optional
ESTEL analog inputs/outputs, serial link: 1 RS-232/422; USB port; Ethernet port (TCP/IP). Instrument must be operated in
accordance with the Ambient Air Continuous Particulate Monitor Model MP 101M operation manual. This designation applies
to PM2 5 measurements only.

Federal Register: Vol. 78, page 67360,11/12/2013

Graseby Andersen Model RAAS2.5-100 PM2.5 Ambient Air Sampler
Manual Reference Method: RFPS-0598-119

"Graseby Andersen Model RAAS2.5-100 PM2 5 Ambient Air Sampler," operated with software version 4B, 5.0.1 - 6.09, 6.OA,
or 6.0B, configured for "Single 2.5" operation, for a continuous 24-hour sample period at a flow rate of 16.67 liters/minute,
and in accordance with the Model RAAS2.5-100 Operator's Manual and with the requirements and sample collection filters
specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 63, page 31991, 06/11/1998


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 12

Graseby Andersen Model RAAS2.5-300 PM2.5 Sequential Ambient Air Sampler
Manual Reference Method: RFPS-0598-120

"Graseby Andersen Model RAAS2.5-300 PM2 5 Sequential Ambient Air Sampler," operated with software version 4B, 5.0.1 -
6.09, 6.0A, or 6.0B, configured for "Multi 2.5" operation, for a continuous 24-hour sample period at a flow rate of 16.67
liters/minute, and in accordance with the Model RAAS2.5-300 Operator's Manual and with the requirements and sample
collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 63, page 31991, 06/11/1998

Grimm Model EDM 180 PM2. s Monitor
Automated Equivalent Method: EQPM-0311-195

"Grimm Technologies, Inc. Model EDM 180 PM2 5 or Tisch Environmental TE-EDM 180 PM2 5 Monitor," light scattering
continuous ambient particulate monitor operated for 24 hours at a volumetric flow rate of 1.2 L/min, configured with a Nafion®-
type air sample dryer, complete for operation with firmware version 7.80, in accordance with the Grimm Technologies, Inc.
Model EDM 180 Operation and Instruction Manual. The optional graphic presentation can be made with the software model
1.177 version 3.30.

Federal Register: Vol. 76, page 15974, 03/22/2011
Latest Modification: 01/2012; 3/2014; 12/2016

Met One Instruments, Inc. BAM-1020 Monitor - PMzs FEM Configuration, Horiba APDA-371- PMzs Configuration, or

Ecotech Spirant BAM1100

Automated Equivalent Method: EQPM-0308-170

"Met One Instruments, Inc. BAM-1020 Beta Attenuation Mass Monitor - PM25 FEM Configuration, Horiba Instruments
APDA-371 Beta Attenuation Mass Monitor - PM2 5 FEM Configuration, or Ecotech Spirant BAM1100", configured with the
standard BX-802 EPA PM10 inlet (or alternative louvered PM10 inlet meeting 40 CFR 50 Appendix L specifications) and with
a BGI VSCC™ Very Sharp Cut Cyclone or Tisch TE-PM2.5C particle size separator, operated for 24 1-hour average
measurements with firmware revision 3.2.4 or later, with or without an inlet tube extension (BX-823), with or without external
enclosures BX-902 or BX-903 and equipped with BX-596 ambient temperature and barometric pressure combination sensor,
internal BX-961 automatic flow controller operated in Actual (volumetric) flow control mode, the standard BX-802 EPA PM10
inlet head and a BGI Veiy Sharp Cut Cyclone (VSCC™ -A) or Tisch TE-PM2.5C cyclone, BX-827 (110V) or BX-830 (230V)
Smart Inlet Heater, with the heater RH set to 35% and the temperature control set to "off", the 8470-1 revision D or later tape
control transport assembly with close geometry beta source configuration, used with standard glass fiber filter tape, COUNT
TIME parameter set for 8 minutes, the SAMPLE TIME parameter set for 42 minutes, BX-302 zero filter calibration kit required
and with or without BX-970 touch-screen display with USB interface. Instrument must be operated in accordance with the
BAM 1020 Particulate Monitor operation manual, revision F or later, the Horiba APDA-371 Monitor operation manual or the
Ecotech Spirant BAM Manual (rev.A or later), as appropriate, and the VSCC™ or Tisch TE-PM2.5C supplemental manual.
Alternatively, BX-851 (110V) or BX-852 (230V) Dew Point Heating System (DPHS) with BX-597 T/BP/RH sensor may be
substituted for the BX-827 (110V) or BX-830 (230V) Smart Inlet Heater with BX-596 T/BP sensor. DPHS must be operated
in accordance with BAM 1020 Particulate Monitor with DPHS operational manual. DPHS requires firmware V3.10.0 or greater
for the standard BAM-1020 and firmware V4.4.0 with the Touch Screen Display Option (BX-970).

Federal Register: Vol. 73, page 13224, 03/12/2008
Latest modifications: 7/2010; 8/2010; 8/2012; 3/2015; 9/2015

Met One Instruments, Inc. BAM-1020 Beta Attenuation Mass Monitor- PM2.5 FEM Configuration
Automated Equivalent Method: EQPM-0715-266

"Met One Instruments, Inc. BAM-1020 Beta Attenuation Mass Monitor - PM25 FEM Configuration," configured for 24 1-
hour average measurements of PM2 5 by beta attenuation, using a glass fiber filter tape roll (460130 or 460180) and a sample
flow rate of 16.67 liters/min and with the standard EPA PM10 inlet (meeting 40 CFR 50 Appendix L specifications) and with a
URG-2000-30EGN PM25 (BX-809) cyclonic separator, BX-596 combo T/RH sensor, BX-827 (110V) or BX-830 (230 V).
Instrument must be operated in accordance with the BAM 1020 Particulate Monitor operation manual, revision L or later.
Instrument must be operated with firmware revision 3.2.4 or later. This PM2.5 equivalent method designation only applies to
the BAM-1020 configured with the URG-2000-30EGN cyclone.

Federal Register: Vol. 80, page 51802, 8/26/2015

Met One Instruments, Inc. BAM-1022 Real Time Beta Attenuation Mass Monitor-Outdoor PM2.5 FEM Configuration
Automated Equivalent Method: EQPM-1013-209

"Met One Instruments, Inc. BAM-1022 Beta Attenuation Mass Monitor - Outdoor PM2 5 FEM Configuration," configured for


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 13

24 1-hour average measurements of PM2 5 by beta attenuation, using a glass fiber filter tape roll (460130 or 460180) and a
sample flow rate of 16.67 liters/min and with the standard (BX-802) EPA PMio inlet (meeting 40 CFR 50 Appendix L
specifications) and with a BGI VSCC™ Very Sharp Cut Cyclone (BX-808) or Tisch TE-PM2.5C particle size separator, and
equipped with external enclosure BX-922 and BX-592 ambient temperature sensor or BX-596 ambient temperature/barometric
combination sensor or BX-597 ambient temperature/barometric pressure/relative humidity combination sensor. Instrument
must be operated in accordance with the BAM 1022 Particulate Monitor operation manual, revision 3 or later. This designation
applies to PM2 5 measurements only.

Federal Register: Vol. 78, page 67360,11/12/2013
Latest modifications: 7/2014; 10/2015

Met One Instruments, Inc. E-FRM- PMzs
Manual Reference Method: RFPS-0315-221

"Met One Instruments, Inc. E-FRM," configured for filter sampling of ambient particulate matter using the US EPA PM10 inlet
specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, equipped with either a BGI VSCC™ cyclone or WINS PM25
fractionator, with a flow rate of 16.67 L/min, using 47 mm PTFE membrane filter media, and operating with firmware version
Rl.1.0 and later, and operated in accordance with the Met One E-FRM PM2 5 operating manual.

Federal Register: Vol. 80, page 32114, 6/05/2015

Met One Instruments, Inc. E-FRM- PMzs
Manual Equivalent Method: EQPS-0316-235

"Met One Instruments, Inc. E-FRM," configured for filter sampling of ambient particulate matter using the US EPA PM10 inlet
specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, equipped with a URG-2000-30EGN Cyclone particle size separator,
and operated for a continuous 24-hour sample period at a flow rate of 16.67 liters/minute, using 47 mm PTFE membrane filter
media, and operating with firmware version Rl.1.0 and later, and operated in accordance with the Met One E-FRM PM25
operating manual.

Federal Register: Vol. 81, page 25397, 4/28/2016

Opsis SM200- Dust Monitor
Automated Equivalent Method: EQPM-0812-203

"Opsis SM200- Dust Monitor" configured for PM25 with the US EPA PM10 inlet specified in 40 CFR 50 Appendix L, followed by a BGI Inc.
Veiy Sharp Cut Cyclone (VSCC™) particle size separator, operated for a continuous 24-hour sample period at a total actual flow rate of
16.67 L/min. using 47mm PTFE membrane filters, a TS200 temperature stabilizer and software version 1.04.16 or later, in accordance with the
Opsis SM200 Dust Monitor Operation and Instruction Guide.

Federal Register: Vol. 77, page 55832, 09/11/2012

Rupprecht & Patashnick Partisol "-FRM Model 2000 PMzs Air Sampler
Manual Reference Method: RFPS-0498-117

"Rupprecht & Patashnick Company, Incorporated Partisol®-FRM Model 2000 PM25 Air Sampler," operated with software
versions 1.102 - 1.202, with either R&P-specified machined or molded filter cassettes, with or without the optional insulating
jacket for cold weather operation, for a continuous 24-hour sample period, in accordance with the Model 2000 Instruction
Manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 63, page 18911, 04/16/1998

Rupprecht & Patashnick Partisol® Model 2000 PMz s Audit Sampler
Manual Reference Method: RFPS-0499-129

"Rupprecht & Patashnick Company, Inc. Partisol® Model 2000 PM2 5 Audit Sampler," configured as a PM2 5 reference method
and operated with software (firmware) version 1.2 - 1.202, for a continuous 24-hour sample period at a flow rate of 16.67
liters/minute, in accordance with the Partisol® Model 2000 Operating Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 64, page 19153, 04/19/1999

Rupprecht & Patashnick Partisol® Model 2000 PMz. s Audit Sampler

Manual Reference Method: RFPS-0499-129 or Manual Equivalent Method: EQPM-0202-144

"Rupprecht & Patashnick Co., Inc. Partisol® Model 2000 PM2 5 FEM Audit Sampler," configured with a BGI VSCC™ Very
Sharp Cut Cyclone particle size separator, and operated with software (firmware) version 1.2 - 1.202, for a continuous 24-hour


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 14

sample period at a flow rate of 16.67 liters/minute, in accordance with the Partisol® Model 2000 Operating Manual and VSCC™
supplemental manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 67, page 15567, 04/02/2002

Rupprecht & Patashnick Partisol91-Plus Model 2025 Sequential Air Sampler
Manual Reference Method: RFPS-0498-118

"Rupprecht & Patashnick Company, Incorporated Partisol®-Plus Model 2025 PM25 Sequential Air Sampler," operated with
any software version 1.003 through 1.4.16, with either R&P-specified machined or molded filter cassettes, for a continuous 24-
hour sample period, in accordance with the Model 2025 Instruction Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 63, page 18911, 04/16/1998

Teledyne Model 602 BetaPLUS Particle Measurement System or SWAM 5a Dual Channel Monitor
Automatic Equivalent Method: EQPM-0912-204

"Teledyne Model 602 Bcta'"/ ,;s Particle Measurement System" or "SWAM 5a Dual Channel Monitof' configured for 1-hour
measurements of PM2.5 by beta attenuation, on either a single (Line A or B) or both sampling lines (Line A and B)
simultaneously, using 47 mm glass fiber filters, at a sample flow set to 16.67 liters/min and software version 05-02.07.63 or
later and with an inlet system comprised of a PM10 pre-impactor inlet (based on European PM10 inlet design) combined with a
BGI VSCC™ PM2.5 cyclone separator. Operated in accordance with the Teledyne Model 602 BetaPLUS Particle Measurement
System Operation Manual.

Federal Register: Vol. 77, page 60985,10/5/2012

Teledyne Advanced Pollution Instrumentation Model T640 PM Mass Monitor
Automatic Equivalent Method: EQPM-0516-236

"Teledyne Advanced Pollution Instrumentation Model T640 PM mass monitor," continuous ambient particulate monitor
operated at a volumetric flow rate of 5.0 Lpm, equipped with a TAPI 5-Lpm sample inlet (P/N: 081050000), TAPI aerosol
sample conditioner (P/N: 081040000), configured for operation with firmware version 1.0.2.126 or later, and operated in
accordance with the Teledyne Model T640 Operations Manual. This designation applies to PM2 5 measurements only.

Federal Register: Vol. 81, page 45285, 07/13/2016

Teledyne Advanced Pollution Instrumentation Model T640 PM Mass Monitor with 640X Option
Automatic Equivalent Method: EQPM-0516-238

"Teledyne Advanced Pollution Instrumentation Model T640 PM mass monitor with 640X option," continuous ambient
particulate monitor operated at a volumetric flow rate of 16.67 Lpm equipped with the louvered PM10 inlet specified in 40 CFR
50 Appendix L, Figs. L-2 thru L-19, TAPI aerosol sample conditioner (P/N: 081040000), configured for operation with
firmware version 1.0.2.126 or later, in accordance with the Teledyne Model T640 Operations Manual. This designation applies
to PM2.5 measurements only.

Federal Register: Vol. 81, page 45285, 07/13/2016

Thermo Electron Model RAAS2.5-100 PM2.5 Ambient Air Sampler

Manual Reference Method: RFPS-0598-119 or Manual Equivalent Method: EQPM-0804-153

"Thermo Electron Corporation Model RAAS2.5-100 FEM PM2 5 Ambient Air Sampler," configured with a BGI VSCC™ Very
Sharp Cut Cyclone particle size separator and operated with software version 06.0B.00 configured for "Single 2.5" operation,
for a continuous 24-hour sample period at a flow rate of 16.67 liters/minute, in accordance with the Model RAAS2.5-100 FEM
Operator's Manual and VSCC™ supplemental manual, and in accordance with the requirements and sample collection filters
specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 69, page 47924, 08/06/2004

Thermo Electron Model RAAS2.5-200 PM2. s Audit Air Sampler

Manual Reference Method: RFPS-0299-128 or Manual Equivalent Method: EQPM-0804-154

"Thermo Electron Corporation Model RAAS2.5-200 FEM PM2 5 Audit Air Sampler," configured with a BGI VSCC™ Very Sharp Cut
Cyclone particle size separator and operated with software version 06.0B.00, for a continuous 24-hour sample period at a flow rate of
16.67 liters/minute, in accordance with the Model RAAS2.5-200 FEM Operator's Manual and VSCC™ supplemental manual, and in
accordance with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 69, page 47924, 08/06/2004


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 15

Thermo Electron Model RAAS2.5-300 PM2.5 Sequential Ambient Air Sampler

Manual Reference Method: RFPS-0598-120 or Manual Equivalent Method: EQPM-0804-155

"Thermo Electron Corporation Model RAAS2.5-300 FEM PM25 Sequential Ambient Air Sampler," configured with a BGI
VSCC™ Very Sharp Cut Cyclone particle size separator and operated with software version 06.0B.00 configured for "Multi
2.5" operation, for a continuous 24-hour sample period at a flow rate of 16.67 liters/minute, in accordance with the Model
RAAS2.5-300 FEM Operator's Manual and VSCC™ supplemental manual, and in accordance with the requirements and sample
collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 69, page 47924, 08/06/2004

Thermo Environmental Instruments, Incorporated Model 605 "CAPS" Sampler
Manual Reference Method: RFPS-1098-123

"Thermo Environmental Instruments, Incorporated Model 605 "CAPS" Computer Assisted Particle Sampler," configured as a PM2.5
reference method and operated with software version 1.02 A, for a continuous 24-hour sample period, in accordance with the Model
605 Instruction Manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 63, page 58036,10/29/1998

Thermo Scientific Model 5014i or Thermo Scientific FH62C14-DHS Continuous Ambient Particle Monitor
Automated Equivalent Method: EQPM-0609-183

"Thermo Scientific Model 5014i or FH62C14-DHS Continuous Ambient Particle Monitor," operated at a flow rate of 16.67
liters per minute for 24-hour average measurements configured for PM2 5 with a louvered PM10 size selective inlet as specified
in 40 CFR 50 Appendix L, Figs. L-2 through L-19, aPM2.5 BGI Inc. Very Sharp Cut Cyclone (VSCC™) particle size separator,
inlet connector, sample tube, DHS heater with 35% RH threshold, mass foil kit, GF10 filter tape, 8-hour filter change, and
operational calibration and servicing as outlined in the 5014i Continuous Ambient Particulate Monitor or FH62C14-DHS
Continuous Ambient Particulate Monitor operating manual.

Federal Register: Vol. 74, page 28696, 06/17/2009
Latest modification: 03/2010

Thermo Scientific Model 5030i SHARP Monitor or Model 5030 SHARP Monitor
Automated Equivalent Method: EQPM-0609-184

"Thermo Scientific Model 5030 SHARP Monitor," operated at a flow rate of 16.67 liters per minute for 24-hour average
measurements configured for PM2 5 with a louvered PM10 size selective inlet as specified in 40 CFR 50 Appendix L, Figs. L-2
through L-19, a PM2 5 BGI Inc. Very Sharp Cut Cyclone (VSCC™) particle size separator, inlet connector, sample tube, DHS
heater with 58% RH threshold, mass foil kit, GF10 filter tape, nephelometer zeroing kit, 8-hour filter change, and operational
calibration and servicing as outlined in the Model 5030i or Model 5030 SHARP instructional manual.

Federal Register: Vol. 74, page 28696, 06/17/2009
Latest modification: 06/2013

Thermo Scientific TEOM® 1400a Ambient Particulate Monitor with Series 8500C FDMS®; Thermo Scientific TEOM®
1405-F Ambient Particulate Monitor with FDMS*

Automated Equivalent Method: EQPM-0609-181

"Thermo Scientific TEOM® 1400a Ambient Particulate Monitor with Series 8500C FDMS® (Filter Dynamics Measurement
System) or Thermo Scientific TEOM® 1405-F Ambient Particulate Monitor with FDMS®," configured for PM2 5 with the US
EPA PM10 inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, followed by a BGI Inc. Very Sharp Cut Cyclone
(VSCC™) particle size separator, operated with a total actual flow of 16.67 L/min., loaded with Series FDMS® 8500 module
operating software and an FDMS® kit. TEOM® 1400a with Series 8500C FDMS® operated with firmware version 3.20 and
later and TEOM® 1405-F with FDMS® operated with version 1.55 or later and according to the appropriate operating manual.
Federal Register: Vol. 74, page 28696, 06/17/2009
Latest Modification: 09/2010

Thermo Scientific TEOM® 1405-DFDichotomous Ambient Particulate Monitor with FDMS*

Automated Equivalent Method: EQPM-0609-182

"Thermo Scientific TEOM® 1405-DF Dichotomous Ambient Particulate Monitor with FDMS®," configured for dual filter
sampling of fine (PM2 5) and coarse particles using the US EPA PM10 inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru
L-19 and a virtual impactor, with a total flow rate of 16.67 L/min, fine sample flow of 3 L/min, and coarse sample flow rate of
1.67 L/min, and operating with firmware version 1.50 and later, operated with or without external enclosures, and operated in


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 16

accordance with the Thermo Scientific TEOM® 1405-DF Dichotomous Ambient Particulate Monitor Instruction Manual. This
designation applies to PM2 5 measurements only.

Federal Register: Vol. 74, page 28696, 06/17/2009

Thermo Scientific Partisol" 2000-D Dichotomous Air Sampler or Thermo Fisher Scientific Partisol" 2000i-D Dichotomous
Air Sampler

Manual Equivalent Method: EQPS-0509-177

"Thermo Scientific Partisol® 2000-D Dichotomous Air Sampler" or "Thermo Fisher Scientific Partisol® 2000/-D Dichotomous
Air Sampler," configured for dual-filter, single-event sampling of fine (PM2 5) and coarse (PMi 0-2.5) particles, operated with a
U.S. EPA PMi0 inlet and using a virtual impactor to separate fine and coarse PM into two samples for collection on two separate
filter membranes, for a 24-hour sampling period, in accordance with the Partisol® 2000-D or Partisol® 2000/-D instruction
manual, as appropriate. Partisol® 2000/-D operated with firmware version 2.0 or greater.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest modification: 06/2011

Thermo Scientific Partisol-Plus 2025-D Dichotomous Sequential Air Sampler or Thermo Fisher Scientific
Partisol® 2025i-D Dichotomous Sequential Air Sampler
Manual Equivalent Method: EQPS-0509-179

"Thermo Scientific Partisol®-Plus 2025-D Dichotomous Sequential Air Sampler" or "Thermo Fisher Scientific Partisol® 2025/'-
D Dichotomous Sequential Air Sampler," configured for dual-filter sampling of fine (PM25) and coarse (PMi 0-2.5) particle
components, with a U.S. EPA PM10 inlet and using a virtual impactor to separate the fine and coarse PM into two samples for
collection on two separate filter membranes, and operated with the modified filter shuttle mechanism implemented May 31,
2008, and firmware version 1.500 or greater for the Partisol®-Plus 2025-D and version 2.0 or greater for the Partisol® 2025/'-
D, for continuous 24-hour sample period , in accordance with the Partisol®-Plus 2025-D or Partisol® 2025/-D instruction
manual, as appropriate.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest modification: 06/2011

Thermo Scientific Partisol ' 2000-FRM PM2.5 Air Sampler or Thermo Fisher Scientific Partisol® 2000i PM 2.5 Air Sampler
or Rupprecht & Patashnick Partisol" -FRM 2000 PM2. s Air Sampler

Manual Reference Method: RFPS-0498-117 or Manual Equivalent Method: EQPM-0202-143

"Thermo Scientific Partisol® 2000-FRM PM2.5 Air Sampler" or "Thermo Fisher Scientific Partisol® 2000i PM2.5 Air Sampler"
or "Rupprecht & Patashnick Partisol®-FRM 2000 PM2 5 [FEM] Air Sampler configured with a BGI VSCC™ Very Sharp Cut
Cyclone particle size separator with either R&P-specified machined or molded filter cassettes, for 24-hour continuous sampling
periods, in accordance with the Partisol® 2000-FRM or Partisol®2000/' Instruction Manual, as appropriate, and VSCC™
supplemental manual. Partisol® 2000-FRM operated with software versions 1.102 through 1.202 and Partisol® 2000i with
firmware version 2.0 or greater. Method is operated with the requirements and sample collection filters specified in 40 CFR
Part 50, Appendix L, and with or without the optional insulating jacket for cold weather operation.

Federal Register: Vol. 67, page 15567, 04/02/2002
Latest modification: 06/2011

Thermo Scientific Partisol"-Plus 2025 Sequential PM2.5 Air Sampler or Thermo Fisher Scientific Partisol® 2025i Sequential
PMisAir Sampler or Rupprecht & Patashnick Partisol®-Plus 2025 PM2 5 Sequential Sampler
Manual Reference Method: RFPS-0498-118 or Manual Equivalent Method: EQPM-0202-145

"Thermo Scientific Partisol®-Plus 2025 PM2.5 Sequential Air Sampler" or "Thermo Fisher Scientific Partisol® 2025/ PM2.5
Sequential Air Sampler" or "Rupprecht & Patashnick Partisol®-Plus 2025 PM2 5 (FEM) Sequential Air Sampler," configured
with a BGI VSCC™ Very Sharp Cut Cyclone particle size separator with either R&P-specified machined or molded filter
cassettes, for 24-hour continuous sampling periods. Partisol®-Plus 2025 to be operated with any software version 1.003 through
1.5 and Partisol® 2025/ with firmware version 2.0 or greater, and with the modified filter shuttle mechanism. Method to be
operated in accordance with the Partisol®-Plus 2025 or Partisol® 2025/ instruction manual, as appropriate, with the VSCC™
supplemental manual, and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 67, page 15567, 04/02/2002
Latest modification: 06/2011

Tisch Environmental Model TE-Wilbur2.5 Particulate Sampler - PM2 5
Manual Reference Method: RFPS-1014-219

"Tisch Environmental Model TE-Wilbur2.5 PM2.5 Low-Volume Air Particulate Sampler," configured as a PM2.5 reference


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 17

method, operated with software version 3.5 or earlier and a TE-PM10-D PMio size-selective inlet as specified in 40 CFR 50
Appendix L Figs. L-2 thru L-19, with either a BGIVSCC™ Very Sharp Cut Cyclone particle size separator or WINS impactor
or Tisch Environmental TE-PM2.5C cyclone, and operated for 24 sample periods at a flow rate of 16.67 L/min, using 47 mm
PTFE membrane filter media, operated with or without the optional TE-W-600 Solar Panel Power Supply kit, and in accordance
with the Tisch Environmental Model TE-Wilbur2.5 PM2 5 Low-Volume Air Particulate Sampler instruction manual and with
the requirements and sample collection filters as specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 79, page 65392,11/04/2014
Latest modification: 6/2015; 11/2016

Tisch Environmental Model TE-Wilbur2.5 Particulate Sampler - PM2 s
Manual Equivalent Method: EQPS-0415-223

"Tisch Environmental Model TE-Wilbur2.5 PM2 5 Low-Volume Air Particulate Sampler," configured as a PM2 5 equivalent
method, operated with software version 3.5 or earlier and a TE-PM10-D PM10 size-selective inlet as specified in 40 CFR 50
Appendix L Figs. L-2 thru L-19, configured with a Tisch TE-PM2.5C particle size separator and operated for a continuous
24-hour sample period at a flow rate of 16.67 L/min, using 47 mm PTFE membrane filter media, operated with or without the
optional TE-W-600 Solar Panel Power Supply kit, and in accordance with the Tisch Environmental Model TE-Wilbur2.5
PM2 5 Low-Volume Air Particulate Sampler instruction manual and with the requirements and sample collection filters as
specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 80, page32114, 6/05/2015

URG-MASS100 Single PMzs Sampler
Manual Reference Method: RFPS-0400-135

"URG-MASS100 Single PM2 5 FRM Sampler," operated with software (firmware) version 4B or 5.0.1, configured for "Single
2.5" operation with either a BGI Very Sharp Cut Cyclone (VSCC™) particle size separator or WINS impactor, and operated
for a continuous 24-hour sample period at a flow rate of 16.67 liters/minute and in accordance with the URG-MASS 100
Operator's Manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol 65, page 26603, 05/08/2000

URG-MASS300 Sequential PMzs Sampler
Manual Reference Method: RFPS-0400-136

"URG-MASS300 Sequential PM25 FRM Sampler," operated with software (firmware) version 4B or 5.0.1, configured for
"Multi 2.5" operation with either a BGI Very Sharp Cut Cyclone (VSCC™) particle size separator or WINS impactor and
operated for a continuous 24-hour sample period at a flow rate of 16.67 liters/minute, and in accordance with the URG-
MASS300 Operator's Manual and with the requirements and sample collection filters specified in 40 CFR Part 50, Appendix L.
Federal Register: Vol. 65, page 26603, 05/08/2000

URG-MASS100 Single PMzs Sampler
Manual Equivalent Method: EQPS-0316-233

"URG-MASS100 Single PM25 Sampler," operated with software (firmware) version 4B or 5.0.1, configured for "Single 2.5"
operation with a URG-2000-30EGN Cyclone particle size separator, and operated for a continuous 24-hour sample period at a
flow rate of 16.67 liters/minute, and in accordance with the URG-MASS100 Operator's Manual and with the requirements and
sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol. 81, page 25397, 04/28/2016

URG-MASS300 Sequential PMzs Sampler
Manual Equivalent Method: EQPS-0316-234

"URG-MASS300 Sequential PM2 5 Sampler," operated with software (firmware) version 4B or 5.0.1, configured for "Multi
2.5" operation with a URG-2000-30EGN Cyclone particle size separator, and operated a continuous 24-hour sample period at
a flow rate of 16.67 liters/minute, and in accordance with the URG-MASS300 Operator's Manual and with the requirements
and sample collection filters specified in 40 CFR Part 50, Appendix L.

Federal Register: Vol 81, page 25397, 04/28/2016


-------
PMio-2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 18

Particulate Matter - PM10-2.5

BGI Incorporated Model PQ200 Sampler Pair
Manual Reference Method: RFPS-1208-173

"BGI Incorporated Model PQ200 PMi0-2.5 Sampler Pair," for the determination of coarse particulate matter as PMi0-2.5
consisting of a pair of BGI Model PQ200 samplers, with one configured for sampling PM25 (RFPS-0498-116) and the other
configured for sampling PMi0c (RFPS-1298-125) with the PM2 5 separator replaced with a BGI WINS Eliminator and operated
in accordance with the Model PQ200 Instruction manual supplement Appendix O.

Federal Register: Vol. 73, page 77024,12/18/2008

Met One Instruments BAM-1020 PM10-2.5 Measurement System
Automated Equivalent Method: EQPM-0709-185

"Met One Instruments BAM-1020 PM10-2.5 Measurement System," consisting of 2 BAM-1020 monitors, the first of which
(PM25 measurement) is configured as a PM2.5 FEM (EQPM-0308-170). The second BAM-1020 monitor (PM10 measurement)
is configurable as a PM2.5 FEM (EQPM-0308-170), but set to monitor PM10. The BAM-1020 monitors are collocated to within
1-4 meters of one another. The BAM-1020 performing the PM2.5 measurement is equipped with Met One Instruments, Inc. P/N
BX-Coarse interface board and accessories; the units are interconnected to provide concurrent sampling and to report PMi 0-2.5
concentrations directly to the user. Both units are operated in accordance with BAM-1020 PM-Coarse Addendum Rev. 5-5 or
later and the BAM-1020 Operations Manual Rev. D or later.

Federal Register: Vol. 74, page 28241, 06/15/2009

Met One Instruments, Inc. E-FRM- PM10 and E-FRM- PM2.5 Sampler Pair
Manual Reference Method: RFPS-0316-232

"Met One Instruments, Inc. E-FRM- PM10 and E-FRM- PM2.5 Sampler Pair" for the determination of coarse particulate
matter as PMi 0-2.5, consisting of a pair of Met One Instruments, Inc. E-FRM samplers, with one being the E-FRM PM2 5
sampler (RFPS-0315-221) and the other being the E-FRM PM10 sampler (RFPS-0216-231). The units are collocated to within
1-4 meters of one another. Both units are operated in accordance with the associated E-FRM instruction manual. This
designation applies to PMi 0-2.5 measurements only.

Federal Register: Vol. 81, page 25397, 04/28/2016

Teledyne Model 602 BetaPLUS Particle Measurement System or SWAM 5a Dual Channel Monitor
Automated Equivalent Method: EQPM -0912-206

"Teledyne Model 602 Bcta'"/ , :s Particle Measurement System" or "SWAM 5a Dual Channel Monitor" configured for 1-hour
measurements of PM10 and PM2.5 by beta attenuation, with the standard, louvered US EPA PM10 size selective inlet specified
in 40 CFR 50 Appendix L on one channel (Line A or B) and with an inlet system comprised of a PM10 pre-impactor inlet (based
on European PM10 inlet design) combined with a BGI VSCC™ PM2 5 cyclone separator on the second channel (Line A or B,
but always with PM10 on the opposite Line). The PMi0-2.5 mass measurement is performed using the resultant subtraction of
PM10 minus PM2 5. Operated in accordance with the Teledyne Model 602 Bcta'"/ , :s Particle Measurement System Operation
Manual.

Federal Register: Vol. 77, page 60985,10/5/2012

Teledyne Advanced Pollution Instrumentation Model T640 PM Mass Monitor with 640X Option
Automated Equivalent Method: EQPM-0516-240

"Teledyne Advanced Pollution Instrumentation Model T640 PM mass monitor with 640X option," continuous ambient
particulate monitor operated at a volumetric flow rate of 16.67 Lpm, equipped with the louvered PM10 inlet specified in 40
CFR 50 Appendix L, Figs. L-2 thru L-19, TAPI aerosol sample conditioner (P/N: 081040000), configured for operation with
firmware version 1.0.2.126 or later, in accordance with the Teledyne Model T640 Operations Manual. This designation
applies to PMi 0-2.5 measurements only.

Federal Register: Vol. 81, page 45285, 07/13/2016

Thermo Scientific Partisol" 2000-D Dichotomous Air Sampler or Thermo Fisher Scientific Partisol" 2000i-D Dichotomous
Air Sampler

Manual Equivalent Method: EQPS-0509-178

"Thermo Scientific Partisol® 2000-D Dichotomous Air Sampler" or "Thermo Fisher Scientific Partisol® 2000/-D Dichotomous
Air Sampler," configured for dual-filter, single-event sampling of fine (PM2.5) and coarse (PMi 0-2.5) particles, operated with a


-------
PMio-2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 19

U.S. EPA PMio inlet and using a virtual impactor to separate fine and coarse PM into two samples for collection on two separate
filter membranes, for a 24-hour sampling period, in accordance with the Partisol® 2000-D or Partisol® 2000/-D instruction
manual, as appropriate. Partisol® 2000/-D operated with firmware version 2.0 or greater.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest modification: 06/2011

Thermo Scientific Partisol-Plus 2025-D Dichotomous Sequential Air Sampler or Thermo Fisher Scientific Partisol9 2025i-
D Dichotomous Sequential Air Sampler
Manual Equivalent Method: EQPS-0509-180

"Thermo Scientific Partisol®-Plus 2025-D Dichotomous Sequential Air Sampler" or "Thermo Scientific Partisol® 2025/-D
Dichotomous Sequential Air Sampler," configured for dual-filter sampling of fine (PM25) and coarse (PMi 0-2.5) particle
components, with a U.S. EPA PM10 inlet and using a virtual impactor to separate the fine and coarse PM into two samples for
collection on two separate filter membranes, and operated with the modified filter shuttle mechanism implemented May 31,
2008, and firmware version 1.500 or greater for the Partisol®-Plus 2025-D and version 2.0 or greater for the Partisol® 2025/'-
D, for 24-hour continuous sampling periods, in accordance with the Partisol®-Plus 2025-D or Partisol® 2025/-D instruction
manual, as appropriate.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest modification: 06/2011

Thermo Scientific Partisol® 2000-FRM PM10-2.5 Sampler Pair or Thermo Fisher Scientific Partisol!® 2000i PM10-2.5 Air
Sampler Pair

Manual Reference Method: RFPS-0509-175

"Thermo Scientific Partisol® 2000-FRM PM10-2.5 Sampler Pair" or "Thermo Fisher Scientific Partisol® 2000/ PMi 0-2.5 Air
Sampler Pair," for the determination of coarse particulate matter as PMi 0-2.5, consisting of a pair of Thermo Scientific Partisol®
2000-FRM or 2000/ samplers, with one configured as a PM2 5 sampler (RFPS-0498-117) and the other configured as a PMioc
sampler with the PM2 5 separator replaced with a Thermo Scientific WINS bypass downtube (RFPS-1298-126), with U.S. EPA
PM10 inlets on both samplers and operated in accordance with the Partisol® 2000-FRM or 2000/ instruction manual supplement,
as appropriate, and the 2000/ operated with firmware version 2.0 or greater.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest modification: 06/2011

Thermo Scientific Partisol"-Plus 2025 Sequential PM10-2.5 Air Sampler Pair or Thermo Fisher Scientific Partisol® 2025i
Sequential PMio-is Air Sampler Pair
Manual Reference Method: RFPS-0509-176

"Thermo Scientific Partisol®-Plus 2025 Sequential PM10-2.5 Air Sampler Pair" or "Thermo Fisher Scientific Partisol® 2025/
Sequential PMi 0-2.5 Air Sampler Pair," for the determination of coarse particulate matter as PM10-2.5, consisting of a pair of Thermo
Scientific Partisol®-Plus 2025 sequential samplers or a pair of Thermo Fisher Scientific Partisol® 2025/ sequential samplers, with
one configured as a PM2 5 sampler (RFPS-0498-118) and the other configured as a PMi0c sampler with the PM2 5 separator replaced
with a Thermo Scientific Partisol® 2025 downtube (RFPS-1298-127). Partisol®-Plus 2025 to be operated with any software version
1.003 through 1.5 and Partisol® 2025/ with firmware version 2.0 or greater, with the modified filter shuttle mechanism. Method to
be operated in accordance with the Partisol®-Plus 2025 or Partisol® 2025/ instruction manual supplement, as appropriate.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest modification: 06/2011

Thermo Scientific TEOM® 1405-DFDichotomous Ambient Particulate Monitor with FDMS®

Automated Equivalent Method: EQPM-1013-207

"Thermo Scientific TEOM® 1405-DF Dichotomous Ambient Particulate Monitor with FDMS®," configured for dual filter
sampling of fine (PM2 5) and coarse particles using the US EPA PM10 inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru
L-19 and a virtual impactor, with a total flow rate of 16.67 L/min, fine sample flow of 3 L/min, and coarse sample flow rate of
1.67 L/min, and operating with firmware version 1.70 and later, operated with or without external enclosures, and operated in
accordance with the Thermo Scientific TEOM® 1405-DF Dichotomous Ambient Particulate Monitor Instruction Manual. This
designation applies to PMi 0-2.5 measurements only.

Federal Register: Vol. 78, page 67360,11/12/2013

Tisch Environmental Model TE- Wilbur Low- Volume Air Particulate Sampler Pair
Manual Reference Method: RFPS-1014-220

"Tisch Environmental Model TE-Wilbur Low-Volume Air Particulate Sampler Pair" for the determination of coarse particulate
matter as PMi0-2.5, consisting of a pair of Tisch Environmental Model TE-Wilbur samplers, with one being the TE-Wilbur2.5


-------
PMio-2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 20

PM2.5 sampler (RFPS-1014-219) and the other being a TE-WilburlO PM10 sampler (RFPS-0714-216), and operated in
accordance with the associated TE-Wilbur instruction manual. This designation applies to PMi0-2.5 measurements only.
Federal Register: Vol. 79, page 65392,11/04/2014
Latest modification: 11/2016

Tisch Environmental Model TE-Wilbur2.5 and TE-WilburlO Sampler Pair
Manual Equivalent Method: EQPS-0415-224

"Tisch Environmental Model TE-Wilbur Low-Volume Air Particulate Sampler Pair" for the determination of coarse particulate
matter as PMi0-2.5, consisting of a pair of Tisch Environmental Model TE-Wilbur samplers, with one being the TE-Wilbur2.5
PM2.5 sampler with TE-PM2.5C particle size separator (EQPS-0415-223) and the other being a TE-WilburlO PM10 sampler
(RFPS-0714-216), and operated in accordance with the associated TE-Wilbur instruction manual. This designation applies to
PM10-25 measurements only.

Federal Register: Vol 80, page 32114, 6/05/2015


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016

Page 21

Sulfur Dioxide

Reference Method for S02 (Pararosaniline Method)

Manual Reference Method: 40 CFR Part 50, Appendix A

Reference Method for the Determination of Sulfur Dioxide in the Atmosphere (Pararosaniline Method)

Federal Register: Vol. 47, page 54899,12/06/1982 and Vol 48,17355, 04/22/1983

Pararosaniline Method for S02 - Technicon I
Manual Equivalent Method: EQS-0775-001

"Pararosaniline Method for the Determination of Sulfur Dioxide in the Atmosphere-Technicon I Automated Analysis System"
Federal Register: Vol. 40, page 34024, 08/13/1975

Pararosaniline Method for S02 - Technicon II
Manual Equivalent Method: EQS-0775-002

"Pararosaniline Method for the Determination of Sulfur Dioxide in the Atmosphere-Technicon II Automated Analysis System"
Federal Register: Vol. 40, page 34024, 08/13/1975

Advanced Pollution Instrumentation, Inc. Model 100 S02 Analyzer
Automated Equivalent Method: EQSA-0990-077

"Advanced Pollution Instrumentation, Inc. Model 100 Fluorescent SO2 Analyzer," operated on the 0-0.1 ppm1, the 0-0.2 ppm1,
the 0-0.5 ppm, or the 0-1.0 ppm range with a 5-micron TFE filter element installed in the rear-panel filter assembly, either a
user- or vendor-supplied vacuum pump capable of providing 20 inches of mercury vacuum at 2.5 L/min, with or without any
of the following options: Internal Zero/Span; Pump Pack; Rack Mount With Slides; RS-232 Interface; Status Output; TFE
Zero/Span Valves; Zero Air Scrubber; Dual Range.2
Federal Register: Vol. 55, page 38149, 09/17/1990

ASARCO Model 500 S02 Monitor
Automated Equivalent Method: EQSA-0877-024

"ASARCO Model 500 Sulfur Dioxide Monitor," operated on a 0-0.5 ppm range; or "ASARCO Model 600 Sulfur Dioxide Monitor,"
operated on a 0-1.0 ppm range. (Both models are identical except for the range.) NOTE: This method is no longer commercially available.
Federal Register: Vol. 42, page 44264, 09/02/77 and Vol 44, page 67522,11/26/1979

Beckman Model 953 Fluorescent Ambient S02 Analyzer
Automated Equivalent Method: EQSA-0678-029

"Beckman Model 953 Fluorescent Ambient SO2 Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with a time
constant setting of 2, 2.5, or 3 minutes, a 5 to 10 micron membrane filter element installed in the rear-panel filter assembly,
with or without any of the following options: Remote Operation Kit, Catalog No. 641984; Digital Panel Meter, Catalog No.
641710; Rack Mount Kit, Catalog No. 641709; Panel Mount Kit, Catalog No. 641708.

Federal Register: Vol. 43, page 35995, 08/14/1978

Bendix Model 8303 Sulfur Analyzer
Automated Equivalent Method: EQSA-1078-030

"Bendix Model 8303 Sulfur Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with a Teflon filter installed on the
sample inlet of the H2S scrubber assembly.

Federal Register: Vol. 43, page 50733,10/31/1978

Columbia Scientific Industries Model 5700 SO2 Analyzer
Automated Equivalent Method: EQSA-0494-095

"Columbia Scientific Industries Model 5700 Sulfur Dioxide Analyzer," operated with software version 1.0 on any full scale
range between 0-250 ppb1 and 0-1000 ppb, at any integration time setting from 20 to 99 seconds, at any temperature in the
range of 15°C to 30°C, at any AC line voltage in the range of 105 to 130 volts, and with or without any of the following options:
964-0121 Alarm Relay Contacts	964-0125 Dual Current Outputs	964-0131 Rack Mount

964-0122 Input Solenoids	964-0126 Printer	964-0012 Single Headed Pump

964-0124 Dual Analog Voltage Outputs
Federal Register: Vol. 59, page 18818, 04/20/1994


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016	Page 22

Dasibi Model 4108 U. V. Fluorescence SO2 Analyzer
Automated Equivalent Method: EQSA-1086-061

"Dasibi Model 4108 U.V. Fluorescence SO2 Analyzer," operated with a range of 0-100 ppb1, 0-200 ppb1, 0-500 ppb, or 0-1000
ppb, with a Teflon-coated particulate filter and continuous hydrocarbon removal system, with or without any of the following
options: Rack Mounting Brackets and Slides; RS-232-C Interface; Temperature Correction.

Federal Register: Vol. 51, page 32244, 09/10/1986

DKK-TOA Corp. Model GFS-32 U.V. FluorescentSO2Analyzer
Automated Equivalent Method: EQSA-0701-115

"DKK-TOA Corporation Model GFS-32 Ambient Air SO2 Ultraviolet Fluorescent Analyzer," operated within the 0.000 to
0.500 ppm range in the temperature range of 20°C to 30°C.

Federal Register: Vol. 62, page 44007, 08/18/1997

DKK-TOA Corp. Model GFS-112E U. V. Fluorescent SO2 Analyzer
Automated Equivalent Method: EQSA-0100-133

"DKK-TOA Corporation Models GFS-112E and GFS-112E-1 U.V. Fluorescence SO2 Analyzer," operated at any temperature
ranging from 15°C to 35°C, on any of the following measurement ranges: 0-0.05 ppm1, 0-0.100 ppm1, 0-0.200 ppm1, 0-0.5
ppm, or 0-1.000 ppm, and with or without the optional Internal zero air supply and permeation tube oven.2
Federal Register: Vol. 65, page 2610, 01/18/2000

DKK-TOA Corp. Model GFS-312E Ambient SO2 Analyzer
Automated Equivalent Method: EQSA-1107-168

"DKK-TOA Corporation Model GFS-312E Ambient SO2 Analyzer," operated at any environmental temperature in the range
of 20°C to 30°C on any of the following measurement ranges1: 0-0.1 ppm, 0-0.2 ppm and 0-0.5 ppm.2
Federal Register: Vol. 72, page 63176,11/08/2007

Ecotech Serinus 50 Sulfur Dioxide Analyzer or Opsis AB OPS50 Sulfur Dioxide Analyzer or Teledyne Analytical
Instruments 6400E Sulfur Dioxide Analyzer or Tisch Environmental TE 2.0 Sulfur Dioxide Analyzer
Automated Equivalent Method: EQSA-0809-188

"Ecotech Serinus 50 Sulfur Dioxide Analyzer" or "Opsis AB OPS50 Sulfur Dioxide Analyzer" or "Teledyne Analytical
Instruments 6400E Sulfur Dioxide Analyzer" or "Tisch Environmental TE 2.0 Sulfur Dioxide Analyzer," operated in the
range of 0-0.5 ppm, with a five-micron Teflon® filter element installed, and with the following selected: Background-Enabled,
Control Loop-Enabled, Diagnostic Mode-Operate, Pres/Temp/Flow Compensation-Enabled, Span Compensation-Disabled,
with concentration automatically corrected for temperature and pressure changes, and operated according to the Serinus 50
Sulfur Dioxide Analyzer User Manual or the Opsis OPS50 Sulfur Dioxide Analyzer User Manual or the Teledyne Analytical
Instruments 6400E Sulfur Dioxide Analyzer Instruction Manual or the Tisch Environmental TE 2.0 Sulfur Dioxide Analyzer
Instruction Manual, as appropriate.

Federal Register: Vol. 74, page 38184, 07/31/2009
Latest Modifications: 05/2010, 05/2011, 05/2012, 8/2014

Environnement S.A. Model AF21M SO2 Analyzer
Automated Equivalent Method: EQSA-0292-084

"Environnement S.A Model AF21M Sulfur Dioxide Analyzer," operated on a range of 0-0.5 ppm with a response time
coefficient setting of 01, a Teflon filter installed in the rear-panel filter assembly, and with or without any of the following
options: Rack Mount/Slides; RS-232-C Interface.

Federal Register: Vol. 57, page 5444, 02/14/1992


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016

Page 23

Environnement S.A. Model AF22M or AF22e SO2 Analyzer
Automated Equivalent Method: EQSA-0802-149

"Environnement S. A Model AF22M or AF22eUV Fluorescence Sulfur Dioxide Analyzer," operated with a Ml scale range of
0 - 500 ppb, at any temperature in the range of 10°C to 35°C, with a 5-micron PTFE sample particulate filter, with automatic
response time, with the automatic "ZERO-REF" cycle ON and set for activation every 24 hours, and with or without either of
the following options: Permeation oven, ESTEL analog input/output board and LCD color touch screen.2
Federal Register: Vol. 67, page 57811, 09/12/2002

Environnement S.A. SANOA Multigas Longpath Monitoring System
Automated Equivalent Method: EQSA-0400-138

"Environnement S.A. Model SANOA Multigas Longpath Air Quality Monitoring System," consisting of a receiver, one or
more projectors, interface unit, a user-provided control unit computer running the SANOA VisionAIR software, and associated
incidental equipment; configured for measuring SO2, with the temperature control and internal calibration cell options installed,
operated with a measurement range of 0 to 0.5 ppm, over an installed monitoring path length of between 27 and 500 meters,
within an ambient air temperature range of -30 to +45°C, with a measurement (integrating) time of 180 seconds, and with or
without external temperature and barometric pressure sensors or any of the following options: external (meteo) input
connection, series 1M bus connection, OGR type projector, analog outputs.

Federal Register: Vol. 65, page 26603, 05/08/2000

Horiba Models APSA-360, APSA-360-CE, or APSA-360A-CE SO2 Monitors
Automated Equivalent Method: EQSA-0197-114

"Horiba Instruments, Inc. Models APSA-360, APSA-360-CE or APSA-360A-CE Ambient Sulfur Dioxide Monitor," operated
with a full scale range of 0 - 0.50 ppm, at any temperature in the range of 5°C to 40°C, with a Line Setting of "MEASURE,"
an Analog Output Setting of "MOMENTARY VALUE", and with or without any of the following options:21) Rack Mounting
Plate and Side Rails, 2) RS-232 Communications Port, and 3) Internal zero gas and span gas generator.

"Horiba Instruments, Inc. Model APSA-360A-CE Ambient Sulfur Dioxide Monitor," operated with one of the following measurement
ranges: 0-0.05 ppm, 0-0.1 ppm, 0-0.2 ppm, 0-0.5 ppm or 0-1.0 ppm; with selectable time constants from 10 to 300 seconds; at any
temperature in the range of 5°C to 40°C; and with or without the optional internal zero gas and span gas generator.

Federal Register: Vol. 62, page 6968, 02/14/97; Vol 63, page 31992, 06/11/1998

Horiba Model APSA-3 70 Ambient SO 2 Monitor
Automated Equivalent Method: EQSA-0506-159

"Horiba Instruments Incorporated Model APSA-370 Ambient SO2 Monitor," operated with a full scale fixed measurement
range of 0 - 0.50 ppm, with the automatic range switching off, at any environmental temperature in the range of 20°C to 30°C.2
Federal Register: Vol. 71, page 25587, 05/01/2006

Lear Siegler Model AM2020 SO2 Monitor
Automated Equivalent Method: EQSA-0486-049

"Lear Siegler Model AM2020 Ambient SO2 Monitor," operated on a range of either 0-0.5 or 0-1.0 ppm, at a wavelength of
299.5 nm, with a 5 minute integration period, over any 10°C temperature range between 20°C and 45°C, with or without the
automatic zero and span correction feature.

Federal Register: Vol. 45, page 79574,12/01/1980; Vol 46, page 9997, 01/30/1981

Lear Siegler Model SMI000 SO2 Monitor
Automated Equivalent Method: EQSA-1275-005

"Lear Siegler Model SM1000 SO2 Ambient Monitor," operated on the 0-0.5 ppm range, at a wavelength of 299.5 nm, with the
"slow" (300 second) response time, with or without any of the following options: SM-1 Internal Zero/Span; SM-2 Span Timer
Card; SM-3 0-0.1 Volt Output; SM-4 0-5 Volt Output; SM-5 Alternate Sample Pump; SM-6 Outdoor Enclosure.

Federal Register: Vol 41, page 3893, 01/27/1976; Vol 41, page 32946, 08/06/1976; Vol. 42, page 13044, 03/08/1977; Vol.
45, page 1147, 01/04/1980


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016	Page 24

Meloy Model SA185-2A SO2 Analyzer
Automated Equivalent Method: EQSA-1275-006

"Meloy Model SA185-2A Sulfur Dioxide Analyzer," operated on the 0-0.5 ppm range, with or without any of the following
options:

S-2 Modified Recorder Output
S-5 Teflon-Coated Block
S-6A Reignite Timer Circuit
S-l 1A Manual Zero and Span
S-35 Front Panel Digital Meter with BCD Output
S-l3 Status Lights	S-22A Remote Zero/Span Control

S-14 Output Booster Amplifier	S-23 Automatic Zero Adjust

S-14B Line Transmitter Board	S-23A Automatic/Manual Zero Adjust

or operated on the 0-1.0 ppm range with either option S-36 or options S-l and S-24, with or without any of the other options.
Federal Register: Vol. 41, page 3893, 01/27/1976 and Vol 43, page 38088, 08/25/1978

S-l Linearized Output

S-24 Dual Range Linearized Output

S-33 Remote Range Control and Status

S-7 Press to Read Volt MeterS-34 Remote Control

S-22 Remote Zero/Span Control and Status (Timer)

S-l IB Automatic Zero and Span

S-36 Dual Range Log-Linear Output

S-38 Sampling Mode Status

S-l8 Rack Mount Conversion
S-18A Rack Mount Conversion
S-21 Front Panel Digital (Signals)

Meloy Model SA285E SO2 Analyzer
Automated Equivalent Method: EQSA-1078-032

"Meloy Model SA285E Sulfur Dioxide Analyzer," operated on the following ranges and time constant switch positions:
Range, ppb:	0-501	0-1001	0-500	0-1000

Time Constant Setting: 1 or 10	1 or 10	off, 1 or 10 off, 1 or 10

The analyzer may be operated at temperatures between 10°C and 40°C and at line voltages between 105 and 130 volts, with or
without any of the following options:

S-5 Teflon Coated Block
S-14B Line Transmitter Board
S-l8 Rack Mount Conversion
S-18A Rack Mount Conversion
S-21 Front Panel Digital Meter
S-22 Remote Zero/Span Control
and Status (Timer)
S-22A Remote Zero/Span Control

S-22B Remote Zero/Span Control

and Status (Pulse)
S-23 Auto Zero Adjust
S-23A Auto/Manual Zero Adjust
S-25 Press To Read
S-26 Manual Zero and Span
S-27 Auto Manual Zero/Span
S-28 Auto Range and Status

S-30 Auto Reignite

S-32 Remote Range Control and Status
S-35 Front Panel Digital Meter With

BCD Output
S-37 Temperature Status Lights
S-38 Sampling Mode Status

Federal Register: Vol. 43, page 50733,10/31/1978

Meloy Model SA 700 Fluorescence Sulfur Dioxide Analyzer
Automated Equivalent Method: EQSA-0580-046

"Meloy Model SA 700 Fluorescence Sulfur Dioxide Analyzer," operated on the 0-250 ppb1, the 0-500 ppb, or the 0-1000 ppb
range with a time constant switch position of either 2 or 3. The analyzer may be operated at temperatures between 20°C and
30°C and at line voltages between 105 and 130 volts, with or without any of the following options: FS-1 Current Output; FS-2
Rack Mount Conversion; FS-2A Rack Mount Conversion; FS-2B Rack Mount Conversion; FS-3 Front Panel Mounted Digital
Meter; FS-5 Auto/Manual Zero/Span With Status; FS-6 Remote/Manual Zero/Span With Status; FS-7 Auto Zero Adjust.
Federal Register: Vol. 45, page 31488, 05/13/1980

Monitor Labs Model 8450 Sulfur Monitor
Automated Equivalent Method: EQSA-0876-013

"Monitor Labs Model 8450 Sulfur Monitor," operated on a range of either 0-0.5 or 0-1.0 ppm, with a 5 second time constant,
a model 8740 hydrogen sulfide scrubber in the sample line, with or without any of the following options: BP Bipolar Signal
Processor; IZS Internal Zero/Span Module; V Zero/Span Valves; CLO Current Loop Output; TF TFE Sample Particulate Filter;
VT Zero/Span Valves and Timer; DO Status Remote Interface.

Federal Register: Vol. 41, page 36245, 08/27/1976 and Vol 44, page 33476, 06/11/1979

Monitor Labs/Lear Siegler Model 8850 SO2 Analyzer
Automated Equivalent Method: EQSA-0779-039

"Monitor Labs or Lear Siegler Model 8850 Fluorescent SO2 Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with
an internal time constant setting of 55 seconds, a TFE sample filter installed on the sample inlet line, with or without any of the
following options: 03A Rack; 03B Slides; 05A Valves Zero/Span; 06A IZS Internal Zero/Span Source; 06B,C,D NIST-
Traceable Permeation Tubes; 08A Pump; 09A Rack Mount For Option 08A; 010 Status Output W/Connector; 013 Recorder
Output Options; 014 DAS Output Options; 017 Low Flow Option; 018 Kicker.

Federal Register: Vol. 44, page 44616, 07/30/1979


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016	Page 25

Monitor Labs/Lear Siegler Model 8850S SO2 Analyzer
Automated Equivalent Method: EQSA-0390-075

"Monitor Labs or Lear Siegler Model 8850S SO2 Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm.

Federal Register: Vol. 55, page 5264, 02/14/1990

Opsis Model AR 500 and System 300 Open Path Ambient Air Monitoring Systems for SO2
Automated Equivalent Method: EQSA-0495-101

"Opsis Model AR 500 System" or "System 300 Open Path (long path) Ambient Air Monitoring Systems," configured for
measuring SO2, with one detector and movable grating, operated with a measurement range of 0 to 0.5 ppm or 0 to 1.0 ppm,
an installed monitoring path length between 20 and 500 meters (or 20 and 1000 meters with the ER 150 option, AR 500 System
only), xenon lamp type B (150 watt), fiber optic cable length between 3 to 20 meters; operating within an ambient air
temperature range of -50 to +50°C, an analyzer temperature range of 20 to 30°C, a measurement (integrating) time setting
between 30 and 120 seconds (0 min:30 sec. to 2 min:00 sec.), and with a complete cycle time of not more than 200 seconds (3
min, 20 sec.). Under this method designation, the Model AR 500 System or System 300 consists of: AR 500 opto-analyser;
emitter EM 110 and receiver RE 110 (together identified as ER 110); optic fibre cable OF60-S; power supply PS 150; Opsis
operational software, version 7.0 or 7.1; and initial on-site installation, setup, and limited operator training.2
Optional components that can be used with the Model AR 500 only, in addition to or as alternative to corresponding
components listed above: • AR 503 opto-analyzer configured as Model AR 500 (only the center detector active, sequential
monitoring) • Emitter/receiver ER 150 (for monitoring path lengths up to 1 kilometer) • Transceiver ER 130 and Retroreflector
RE 090 with 7 prisms (max. monitoring path length 150 meters) or 12 prisms (max. monitoring path length 250 meters) • Receiver
RE 130 • Xenon lamp type A (higher short-wavelength UV output) • Optic fibre cable OF60-R (low-loss for short wavelengths)

•	Multiplexers MX 004 and MX 024 • Dataloggers DL 010 and DL 016 • Analogue and digital input/output cards AO 008, AI
016, and DI032 • Analogue and digital isolation cards IA 008, ID 008, OA 008, and OD 008 • Window heaters HF 110 and HF
150 • Mirror heaters HM 110 and HM 150 • Auto calibration unit CU 007 • Software packages IO 80 (for the analogue and
digital input/output adapters), DL10 and DL16 (for data loggers), ComVision, and STAT 500;

Recommended calibration and accuracy audit components (or equivalent) for either Model AR 500 or System 300:

•	Wavelength calibration lamp CA 004 • Calibration bench CB 100 • Receiver unit RE 060 (two required) • Calibration unit
CA 150, with same type lamp as used in the monitoring path emitter • Power supply PS 150 for calibration unit CA 150 •
Calibration cells CC 001-X, where X represents various cell lengths from 1 to 900 mm • Special calibration cells CC 110 or
CC 150 (for mounting directly on receiver) • Light meter LM 010.

Federal Register: Vol 60, page 21518, 05/02/1995

Philips PW9755 SO2 Analyzer

Automated Equivalent Method: EQSA-0676-010

"Philips PW9755 SO2 Analyzer," consisting of the following components: PW9755/02 SO2 Monitor with PW9741/00 SO2
Source, PW9721/00 Filter Set S02, PW9711/00 Electrolyte S02, PW9750/00 Supply Cabinet, PW9750/10 Supply
Unit/Coulometric, either PW9731/00 Sampler or PW9731/20 Dust Filter (or vendor-approved alternate particulate filter);
operated with a 0-0.5 ppm range and with a reference voltage setting of 760 millivolts; with or without any of the following
options: PW9750/30 Frame For MTT; PW9752/00 Air Sampler Manifold; PW9753/00 Mounting Rack For Accessories;
PW9750/41 Control Clock 60 Hz; PW9754/00 Air Distributor.

Federal Register: Vol. 41, page 26252, 6/25/1976; Vol 41, page 46019,10/19/1976; Vol 42, page 28571, 06/03/1977

Philips PW9700 SO2 Analyzer

Automated Equivalent Method: EQSA-0876-011

'Philips PW9700 SO2 Analyzer,' consisting of the following components: PW9710/00 Chemical Unit with PW9711/00,
Electrolyte S02, PW9721/00 Filter Set S02, PW9740/00 S02 Source; PW9720/00 Electrical Unit; PW9730/00 Sampler Unit (or
vendor-approved alternate particulate filter); operated with a 0-0.5 ppm range and with a reference voltage of 760 millivolts.
Federal Register: Vol. 41, page 34105, 08/12/1976

SERES Model SF 2000 G Sulfur Dioxide Analyzer
Automated Equivalent Method: EQSA-0810-194

"SERES model SF 2000 G Sulfur Dioxide Analyzer," UV fluorescence method using a wavelength source approaching 215
nm and a selective membrane for aromatic hydrocarbon removal, operated with a full scale measurement range of 0 - 0.5
ppm at any ambient temperature in the range of 20°C to 30°C, with tabletop or rack mounts, microprocessor controlled
menu-driven user interface, onboard diagnostics and system test functions, analog output signals of 4-20 mA or user


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016

Page 26

selectable voltage ranges up to 10 V, printer port, modem port and 32 pin data/control/alarm port, user selectable manual and
automatic zero/span and calibrate modes; with or without a permeation tube system (optional equipment) for internal
calibration; operated in accordance with the SF 2000 G User and Maintenance Manual.

Federal Register: Vol. 75, page 51039, 08/18/2010

SIR S.A. Model S-5001 U. V. Fluorescence SO2 Analyzer
Automated Equivalent Method: EQSA-0507-166

"SIR S.A. Model S-5001 U.V. Fluorescence SO2 Analyzer," operated with a full-scale measurement range of 0 - 0.5 ppm, with an
integration time setting of 1 minute, and with or without an optional PCMCIA card or the optional internal permeation oven.2
Federal Register: Vol. 72, page 26627, 05/10/2007

Sutron Model 6020 Sulfur Dioxide Fluorescent Analyzer
Automated Reference Method: RFSA-0616-237

"Sutron Model 6020 Sulfur Dioxide Fluorescent Analyzer," operated at any of the following measurement ranges: 0-0.5 ppm,
at any ambient temperature in the range of 5-40 °C, at any line voltage in the range of 90-260 VAC, at any sample flow rate
in the range of 0.4-0.8 L/min, and in accordance with the Model 6020 SO2 Analyzer Operation Manual, with or without the
following options: Zero/span ports for external calibration; an optional inlet filter; or an optional second gas measurement
module collocated inside of the enclosure.

Federal Register: Vol. 81 page 45284, 07/13/2016

Teledyne Advanced Pollution Instrumentation, Models 100A, 100AS, 100E, 100EU, T100, T100U; Teledyne Analytical
Instruments Model 6400A; or Teledyne Monitor Labs sensor-e™ Model TML-50 SO2 Analyzers; or recordum airpointer®
system module 801-001000
Automated Equivalent Method: EQSA-0495-100

'Teledyne Advanced Pollution Instrumentation Models 100A, 100AS, 100E, 100EU, T100 or T100U; Teledyne Analytical
Instruments Model 6400A; or Teledyne Monitor Labs, Inc. sensor-e™ Model TML-50 UV Fluorescent Sulfur Dioxide
Analyzer; operated on any full scale range between 0-50 ppb1 and 0-1000 ppb, at any temperature in the range of 5 to 40 degrees
C, with a TFE filter element or a Kynar® DFU installed in the filter assembly, with either the vendor-supplied internal pump
or a user- or vender-supplied external vacuum pump capable of maintaining an absolute pressure of 35 cm (14 inches) of
mercury (or less) at 1.0 standard liter per minute flow rate, with the following software settings: Dynamic zero: OFF or ON;
Dynamic span: OFF; AutoCal: ON or OFF; Dual range: ON or OFF; Autorange: ON or OFF; Temp/pressure compensation:
ON; dilution factor: OFF or 1.0; and with or without any of the following options (if available for the various models):2 Rack
mount with or without chassis slides; Fluorocarbon zero/span valves; Internal zero/span (IZS); Three-point internal zero/span
(IZS, option 51C); 4-20 mA, isolated analog outputs; analog input option; External pump; Status outputs; Control inputs; Rack
mount for external pump with tray; RS-232 output; Ethernet output; Zero air scrubber; Combustion Filter; SO2 Permeation
tube, certified or uncertified, 0.4 ppm @ 0.7 L/min; SO2 Permeation tube, certified or uncertified, 0.8 ppm @ 0.7 L/min; the
NumaView™ software. Airpointer® module 801-001000 only: operated on any full scale range between 0-0.05 ppm and 0-
1.0 ppm, with a PTFE filter element installed in the internal filter assembly, with the software setting: FRM/FEM conform
mode; at any temperature in the range of 10°C to 45°C, with either a user- or vendor-supplied vacuum pump capable of
providing an absolute pressure of 16 inches mercury or less at 2.5 sLpm; installed in the compact, thermally controlled (-40°C
to + 45°C) and weather proof airpointer® base unit with integrated data acquisition and management system mounted on a
frame, pole, or wall; with or without wireless telemetry; with or without internal span option as module supplement consisting
of permeation oven and permeation tube; with or without modules for other criteria pollutants; with or without analyzer for
particulate matter; with or without additional 3rd party sensors for e.g. meteorology, noise, or traffic counting. Operated with
the appropriate instrument manual. Note 2 applies to the following Teledyne Advanced Pollution Instrumentation Models 100E, 100EU,
T100, T100U, and Teledyne Monitor Labs, Inc. Sensor-e™ Model TML-50.

Federal Register: Vol 60, page 17061, 04/04/1995
Latest Modification: 08/2010; 05/2013; 07/2014; 9/2015

Teledyne Monitor Labs/Casella/Ecotech Model ML9850/CM2050ZEC9850ZEC9850T; Teledyne Monitor Labs/Casella/
Ecotech/Model ML9850B/CM2050B/EC9850B; or Wedding & Associates Model 1040 SO2 Analyzers
Automated Equivalent Method: EQSA-0193-092

"Teledyne Monitor Labs, Casella Monitor, or Ecotech Models ML9850/CM2050/EC9850, orML9850B/CM2050B/EC9850B,
Ecotech Model EC9850T, or Wedding & Associates, Inc. Model 1040 Sulfur Dioxide Analyzers," operated on any full scale


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016

Page 27

range between 0-0.050 ppm1 and 0-1.0 ppm, at any temperature in the range of 15°C to 35°C, with the service switch on the
secondary panel set to the In position; with the following menu choices selected: Range: 0.05 ppm to 1.0ppm; Over-ranging:
Enabled or Disabled; Background: Not Disabled; Calibration: Manual or Timed; Diagnostic Mode: Operate; Filter Type:
Kalman; Pres/Temp/Flow Comp: On; Span Comp: Disabled; and as follows: Model ML9850/CM2050/ EC9850/EC9850T -
with a five-micron Teflon® filter element installed internally, with the 50-pin I/O board installed on the rear panel configured
at any of the following output range settings: Voltage, 0.1 V, 1 V, 5 V, 10 V; Current, 0-20 mA, 2-20 mA, 4-20 mA; and with
or without any of the following options: Valve Assembly for External Zero/Span (EZS); Rack Mount Assembly; Internal
Floppy Disk Drive. Models ML9850B/CM2050B/EC9850B and 1040 - with either a vendor-supplied or equivalent user
supplied five-micron Teflon® filter, zero air scrubber, and exhaust pump, and with or without any of the following options:
Valve Assembly for External Zero/Span (EZS); Rack Mount Assembly; 50-pin I/O board; Exhaust Scrubber; Internal
Zero/Span Assembly (IZS); hinged, fold-down front panel. Operated with the appropriate instrument manual.

Federal Register: Vol. 58, page 6964, 02/03/1993
Latest Modification: 03/2011

Thermo Electron Model 43 SO2 Analyzer
Automated Equivalent Method: EQSA-0276-009

"Thermo Electron Model 43 Pulsed Fluorescent SO2 Analyzer," equipped with an aromatic hydrocarbon cutter and operated on a
range of either 0-0.5 or 0-1.0 ppm, with or without any of the following options: 001 Rack Mounting For Standard 19 Inch Relay
Rack; 002 Automatic Actuation Of Zero and Span Solenoid Valves; 003 Type S Flash Lamp Power Supply; 004 Low Flow.
Federal Register: Vol. 41, page 8531, 02/27/1976; Vol 41, page 15363, 04/12/1976; Vol 42, page 20490, 04/20/1977 Vol.
44, page 21861, 04/12/1979; Vol. 45, page 2700, 01/14/1980; Vol. 45, page 32419, 05/16/1980

Thermo Environmental Instruments, Inc./Thermo Electron Models 43A, 43B, 43C, 43C-TLE, 43i, 43i-TLE SO2 Analyzers
Automated Equivalent Method: EQSA-0486-060

"Thermo Electron or Thermo Environmental Instruments, Inc. Model 43A or 43B Pulsed Fluorescence SO2 Analyzer,"
operated on the 0-0.1 ppm1, the 0-0.2 ppm1, the 0-0.5 ppm, or the 0-1.0 ppm range, with either a high or a low time constant
setting (Model 43 A) and with or without any of the following options:2

001	Teflon Particulate Filter	004 High Flow Rate (1 Lpm)	007Remote Activation of Zero/Span Valves

002	19" Rack Mounting Configuration 005 Current Output	008 RS-232 Interface (Model 43B)

003	Internal Zero/Span Valves	006 Internal Permeation	009 Pressure/Temperature Compensation

Span Source	(Model 43B)

"Thermo Environmental Instruments, Inc. Models 43C, 43C-TLE, 43/, 43i-TLE, Pulsed Fluorescence SO2 Analyzer," operated
on any measurement range between 0-50 ppb1 (0-101 ppb for Model TLE) and 0-1000 ppb, with any time average setting from 10
to 300 seconds, with temperature and/or pressure compensation on or off, operated at temperatures between 20°C and 30°C, and
with or without any of the following options:2 Teflon particulate filter, 4-20 mA current output or I/O expansion board, Rack
mounts, Internal permeation span source/Permeation oven, Internal zero/span and sample solenoid valves, High flow rate (0.5-1.0
Lpm); Models 43C, 43C-TLE: Remote activation of zero/span and sample valves, RS-232/485 interface, airpointer® Model
801-001000-T, operated on any full scale range between 0-0.05 ppm and 0-1.0 ppm, with a PTFE filter element installed in the
internal filter assembly, with the software setting: FRM/FEM conform mode; at any internal temperature in the range of 15°C to
39°C, with either a user- or vendor-supplied vacuum pump capable of providing an absolute pressure of 16 inches mercury or less
at 2.5 sLpm; installed in the compact, thermally controlled (-40°C to + 45°C) and weather proof airpointer® base unit with
integrated data acquisition and management system mounted on a frame, pole, or wall; with or without wireless telemetry; with
or without internal span option as module supplement consisting of permeation oven and permeation tube; with or without modules
for other criteria pollutants; with or without analyzer for particulate matter; with or without additional 3rd party sensors for e.g.
meteorology, noise, or traffic counting. Operated with the appropriate instrument manual.

Federal Register: Vol. 51, page 12390, 04/10/1986
Latest Modification: 10/2015


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 28

Ozone

2B Technologies Model 202 and 205 Ozone Monitors
Automated Equivalent Method: EQOA-0410-190

"2B Technologies Models 202 (single beam) and 205 (dual beam) Ozone Monitors," operated in a range of 0 - 0.5 ppm in an
environment of 10 - 40°C, with temperature and pressure compensation, internal DewLine for humidity control, using a 10
second average, with a 110-220V AC power adapter or a 12 V DC source 4.0 to 6.0 watt power consumption, operated according
to the Model 202 or 205 Ozone Monitor Operation Manual with or without the following: cigarette lighter adapter or a 12V
DC battery for portable operation, external TFE inlet filter and holder, serial data port with computer cable, BNC connector for
0-2.5 V scalable analog output, internal data logger, 3-analog inputs for external signals (such as temperature, relative humidity
or pressure), rack mount hardware, on-board backup sample pump.

Federal Register: VoL75, pages 22126-22127, 04/27/2010
Latest Modification: 12/2010

2B Technologies Model 211 Ozone Monitor
Automated Equivalent Method: EQOA-0514-215

"2B Technologies Models 211 Scrubberless Ozone Monitors," operated in a range of 0 - 0.5 ppm in an environment of 20 -
30°C, with temperature and pressure compensation, internal DewLine for humidity control, gas phase titration of ozone for
interference-free measurements, using a 1 minute average, with a 110-220V AC power adapter or a 12V DC source, 8.0 to 12.0
watt power consumption, operated according to the Model 211 Scrubberless Ozone Monitor Operation Manual with either an
external nitric oxide source or internal photolytic generator for production of NO scrubber gas from nitrous oxide, and with or
without the following: cigarette lighter adapter or a 12V DC battery for portable operation, external PTFE inlet filter and
holder, serial data port with computer cable, BNC connector for 0-2.5 V or 4-20 mA scalable analog output, internal data logger,
microFlash card for data recording and backup, rack mount hardware, and long life sample pump.

Federal Register: Vol.79, pages 34734-34735 06/18/2014

2B Technologies Model 106-L or 106-OEM-L Ozone Monitor
Automated Equivalent Method: EQOA-0914-218

"2B Technologies Model 106-L or 106-OEM-L Ozone Monitor," operated in a range of 0 - 0.5 ppm in an environment of 20-
30 °C, temperature and pressure compensation, internal dewLine for humidity control, using a 1 minute average, with a 12V
DC source supplied by a 100-240V AC power adapter, operated according to the Model 106-L Ozone Monitor Operation
Manual and with or without the following: cigarette lighter adapter or a 12V DC battery for portable operation, external PTFE
or PVDF inlet filter and holder, USB data port with computer cable.

Federal Register: Vol. 79, page 65392,11/04/2014

2B Technologies Model Personal Ozone Monitor (POM)

Automated Equivalent Method: EQOA-0815-227

"2B Technologies Model Personal Ozone Monitor (POM)," operated in a range of 0 - 0.5 ppm in an environment of 20-30 °C,
temperature and pressure compensation, using a 10 second averaging time, with a 12V DC source supplied by a 100-240V AC
power adapter, operated according to the POM Operation Manual and with or without the following: cigarette lighter adapter
or a 12V DC battery or a 7-24 V battery for portable operation, USB data port with computer cable.

Federal Register: Vol. 80, page 51802, 8/26/2015

Beckman Model 950A Ozone Analyzer
Automated Reference Method: RFOA-0577-020

"Beckman Model 950A Ozone Analyzer," operated on a range of 0-0.5 ppm and with the "SLOW" (60 second) response time,
with or without any of the following options: Internal Ozone Generator; Computer Adaptor Kit; Pure Ethylene Accessory.
Federal Register: Vol. 42, page 28571, 06/03/1977

Bendix or Combustion Engineering Model 8002 Ozone Analyzer
Automated Reference Method: RFOA-0176-007

"Bendix or Combustion Engineering Model 8002 Ozone Analyzer," operated on the 0-0.5 ppm range, with a 40 second time
constant, with or without any of the following options: Rack Mounting with Chassis Slides; Rack Mounting without Chassis


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 29

Slides; Zero and Span Timer; Ethylene/CCh Blend Reactant Gas.

Federal Register: Vol. 41, page 5145, 02/04/1976 and Vol 45, page 18474, 03/21/1980

Columbia Scientific Industries Model 2000 Ozone Meter
Automated Reference Method: RFOA-0279-036

"Columbia Scientific Industries Model 2000 Ozone Meter," when operated on the 0-0.5 ppm range with either AC or battery
power: The BCA 952 battery charger/AC adapter M952-0002 (115V) or M952-0003 (230V) is required for AC operation; an
internal battery M952-0006 or 12 volt external battery is required for portable non-AC powered operation.

Federal Register: Vol. 44, page 10429, 02/20/1979

Dasibi Models 1003-AH, 1003-PC, or 1003-RS Ozone Analyzers
Automated Equivalent Method: EQOA-0577-019

"Dasibi Model 1003-AH, 1003-PC, or 1003-RS Ozone Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with or
without any of the following options: Adjustable Alarm; Aluminum Coated Absorption Tubes, Integrated Output; Vycor-
Jacketed U.V. Source Lamp; BCD Digital Output; Rack Mounting Ears and Slides; 0-10 mV, 0-100 mV, 0-1 V, Or 0-10 V;
Glass (Pyrex) Absorption Tubes; Teflon-based Solenoid Valve; Analog Output.

Federal Register: Vol. 42, page 28571, 06/03/1977

Dasibi Models 1008-AH, 1008-PC, or 1008-RS Ozone Analyzers
Automated Equivalent Method: EQOA-0383-056

"Dasibi Model 1008-AH, 1008-PC, or 1008-RS Ozone Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with or without
any of the following options: Aluminum Coated Absorption Tubes; BCD Digital Output; RS232 Interface; Glass (Pyrex) Absorption
Tubes; Vycor-Jacketed U.V. Source Lamp; Ozone Generator; Teflon-based Solenoid Valve; Photometer Flow Restrictor (2 Lpm); 4-
20 mA, Isolated, or Dual Analog Outputs; Rack Mounting Brackets or Slides; 20 Second Update Software.

Federal Register: Vol 48, page 10126, 03/10/1983

DKK-TOA Corp. Model GUX-113E Ozone Analyzer
Automated Equivalent Method: EQOA-0200-134

"DKK-TOA Corporation Models GUX-113E and GUX-113E-1 Ozone Analyzer," operated at any temperature in the range of
15°C to 35°C, on any of the following measurement ranges: 0-0.100 ppm1, 0-0.200 ppm1, 0-0.5 ppm, or 0-1.000 ppm, and with
or without the optional Internal Ozone Generator.2
Federal Register: Vol. 65, page 11308, 03/02/2000

DKK-TOA Corp. Model GUX-313E Ambient O3 Analyzer
Automated Equivalent Method: EQOA-1107-169

"DKK-TOA Corporation Model GUX-313E Ambient O3 Analyzer," operated at any environmental temperature in the range
of 20°C to 30°C on any of the following measurement ranges1: 0-0.1 ppm, 0-0.2 ppm and 0-0.5 ppm.2
Federal Register: Vol. 72, page 63176,11/08/2007

Ecotech Serinus 10 Ozone Analyzer or OpsisAB OPS 10 Ozone Analyzer or Tisch Environmental TE 1.0 Ozone Analyzer
Automated Equivalent Method: EQOA-0809-187

"Ecotech Serinus 10 Ozone Analyzer" or "Opsis AB OPS 10 Ozone Analyzer" or "Tisch Environmental TE 1.0 Ozone
Analyzer," operated in the range of 0-0.5 ppm, with a five-micron Teflon® filter element installed, and with the following
selected: Control Loop-Enabled, Diagnostic Mode-Operate, Pres/Temp/Flow Compensation-Enabled, Span Compensation-
Disabled, with concentration automatically corrected for temperature and pressure changes, and operated according to the
Serinus 10 Ozone Analyzer User Manual, or the Opsis AB OPS 10 Ozone Analyzer User Manual, or the Tisch Environmental
TE 1.0 Ozone Analyzer Instruction Manual, as appropriate.

Federal Register: Vol. 74, page 38184, 07/31/2009
Latest Modifications: 05/2010, 05/2012, 8/2014

Environics Series 300 Ozone Analyzer
Automated Equivalent Method: EQOA-0990-078

"Environics Series 300 Computerized Ozone Analyzer," operated on the 0-0.5 ppm range, with the following parameters
entered into the analyzer's computer system: Absorption Coefficient = 308 V4; Flush Time = 3; Integration Factor = I: Offset


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 30

Adjustment = 0.025ppm; Ozone Average Time = 4; Signal Average = 0; Temp/Press Correction = On; and with or without the

RS-232 Serial Data Interface.

Federal Register: Vol. 55, page 38386, 09/18/1990

Environnement S.A. Model Os41M UV Ozone Analyzer
Automated Equivalent Method: EQOA-0895-105

"Environnement S.A. Model CMIM UV Photometric Ozone Analyzer," operated on a full scale range of 0 - 500 ppb, at any
temperature in the range of 15°C to 35°C, with the response time set to 50 seconds, and with or without any of the following
options:2 Internal Ozone Generator; Span External Control; RS232-422 Serial Interface; Internal Printer.

Federal Register: Vol 60, page 39382, 08/02/1995

Environnement S.A. Model ()t-/2M UV Ozone Analyzer
Automated Equivalent Method: EQOA-0206-148

"Environnement S.A Model 0;,42IVI UV Photometric Ozone Analyzer," operated with a full scale range of 0 - 500 ppb, at any
temperature in the range of 10°C to 35°C, with a 5-micron PTFE sample particulate filter, with response time setting of 11
(Automatic response time), and with or without any of the following options:2 a) Internal ozone generator, b) Span external
control (zero/span solenoid valve).

Federal Register: VoL67, page 42557, 06/24/2002

Environnement S.A. SANOA Multigas Longpath Monitoring System
Automated Equivalent Method: EQOA-0400-137

"Environnement S.A. Model SANOA Multigas Longpath Air Quality Monitoring System, consisting of a receiver, one or more
projectors, interface unit, a user-provided control unit computer running the SANOA VisionAIR software, and associated
incidental equipment; configured for measuring O3, with the temperature control and internal calibration cell options installed,
operated with a measurement range of 0 to 0.5 ppm, over an installed monitoring path length of between 27 and 500 meters,
within an ambient air temperature range of -30 to +45°C, with a measurement (integrating) time of 180 seconds, and with or
without external temperature and barometric pressure sensors or any of the following options: external (meteo) input
connection, series 1M bus connection, OGR type projector, analog outputs. A high-concentration ozone generator, part # 80-
231-03, or the SONIMIX 7121B calibration system is recommended for calibration or accuracy auditing.

Federal Register: Vol.65, page 26603, 05/08/2000

Environnement S.A. Model 03 42e UV Ozone Analyzer
Automated Equivalent Method: EQOA-0515-225

"Environnement S.A. Model 03 42e UV Photometric Ozone Analyzer," operated in a range of 0-0.5 ppm in an environment
of 0-35 °C, with a Teflon sample inlet filter, with automatic temperature and pressure compensation, with a flow-rate of 1 Lpm
and with zero/span external solenoid valve.

Federal Register: Vol. 80, page 32114, 6/05/2015

Horiba Instruments Models APOA-360 or APOA-360-CE Ozone Monitor
Automated Equivalent Method: EQOA-0196-112

"Horiba Instruments, Inc. Model APOA-360 or APOA-360-CE Ambient Ozone Monitor," operated with a full scale range of
0-0.50 ppm, at any temperature in the range of 10°C to 40°C, with a Line Setting of "MEASURE," and an Analog Output of
"MOMENTARY VALUE," and with or without any of the following options:21) Rack Mounting Plate and Side Rails 2) RS-
232 Communications Port, and 3) Optional Internal Zero/Span Check
Federal Register: Vol 61, page 11404, 03/20/1996


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 31

Horiba Instruments Model APOA-3 70 Ozone Monitor
Automated Equivalent Method: EQOA-0506-160

"Horiba Instalments Incorporated APOA-370 Ambient O3 Monitor," standard specification, operated with a full-scale fixed
measurement range of 0 - 0.5 ppm, with the automatic range switching off, at any temperature in the range of 20 to 30°C.2
Federal Register: Vol 71, page 25587, 05/01/2006

McMillan (MEC) Models 1100-1,1100-2, and 1100-3 Ozone Meters
"MEC Model 1100-1 Ozone Meter," Automated Reference Method: RFOA-1076-014
"MEC Model 1100-2 Ozone Meter," Automated Reference Method: RFOA-1076-015
"MEC Model 1100-3 Ozone Meter,' Automated Reference Method: RFOA-1076-016

Operated on a 0-0.5 ppm range, with or without any of the following options: 0011 Rack Mounting Ears; 0026 Alarm Set Feature;
0012 Instrument Bail; 0033 Local-Remote Sample; Zero, Span Kit Blend Feature; 0016 Chassis Slide Kit; 0040 Ethylene/CCh.
Federal Register: Vol. 41, page 46647,10/22/1976; Vol 42, page 30235, 06/13/1977

Meloy Model OA325-2R Ozone Analyzer
Automated Reference Method: RFOA-1075-003

"Meloy Model OA325-2R Ozone Analyzer," operated with a scale range of 0-0.5 ppm, with or without any of the following
options: 0-4 Output Booster Amplifier; 0-18 Rack Mount Conversion; 0-18A Rack Mount Conversion.

Federal Register: Vol. 40, page 54856,11/26/1975

Meloy Model OA350-2R Ozone Analyzer
Automated Reference Method: RFOA-1075-004

"Meloy Model OA350-2R Ozone Analyzer," operated with a scale range of 0-0.5 ppm, with or without any of the following
options: 0-2 Automatic Zero and Span; 0-3 Remote Control Zero and Span; 0-4 Output Booster Amplifier; 0-18 Rack Mount
Conversion; 0-18A Rack Mount Conversion.

Federal Register: Vol. 40, page 54856,11/26/1975

Monitor Labs Model 8410E Ozone Analyzer
Automated Reference Method: RFOA-1176-017

"Monitor Labs Model 8410E Ozone Analyzer," operated on a range of 0-0.5 ppm with a time constant setting of 5 seconds,
with or without any of the following options: DO Status Outputs; ER Ethylene Regulator Assembly; V TFE Zero/Span Valves;
TF TFE Sample Particulate Filter; VT TFE Zero/Span Valves and Timer.

Federal Register: Vol. 41, page 53684,12/08/1976

Monitor Labs/Lear Siegler Model 8810 Ozone Analyzer
Automated Equivalent Method: EQOA-0881-053

"Monitor Labs or Lear Siegler Model 8810 Photometric Ozone Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm,
with selectable electronic time constant settings from 20 through 150 seconds, with or without any of the following options: 05
Pressure Compensation; 06 Averaging Option; 07 Zero/Span Valves; 08 Internal Zero/Span (Valve and Ozone Source); 09
Status; 10 Particulate Filter; 15 through 20 DAS/REC Output.

Federal Register: Vol. 46, page 52224,10/26/1981

Opsis Model AR 500 and System 300 Open Path Ambient Air Monitoring Systems for Ozone
Automated Equivalent Method: EQOA-0495-103

"Opsis Model AR 500 System" or "System 300 Open Path (long path) Ambient Air Monitoring Systems," configured for
measuring O3, with one detector and moveable grating, operated with a measurement range of 0 to 0.5 ppm, an installed
monitoring path length between 20 and 500 meters (or 20 and 1000 meters with the ER 150 option, AR 500 System only),
xenon lamp type B (150 watt), fiber optic cable length between 3 to 20 meters; operating within an ambient air temperature
range of -50 to +50°C, an analyzer temperature range of 20 to 30°C, a measurement (integrating) time setting between 30 and
120 seconds (0 min:30 sec. to 2 min:00 sec.), and with a complete cycle time of not more than 200 seconds (3 min, 20 sec.).
Under this method designation, the Model AR 500 System or System 300 consists of: AR 500 opto-analyser; emitter EM 110
and receiver RE 110 (together identified as ER 110); optic fibre cable OF60-S; power supply PS 150, Opsis operational
software, version 7.0 or 7.1; and initial on-site installation, setup, and limited operator training.2


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 32

Optional components that can be used with the Model AR 500 only, in addition to or as alternative to corresponding
components listed above: • AR 503 opto-analyzer configured as Model AR 500 (only the center detector active, sequential
monitoring) • Emitter/receiver ER 150 (for monitoring path lengths up tol kilometer) • Transceiver ER 130 and Retroreflector
RE 090 with 7 prisms (max. monitoring path length 150 meters) or 12 prisms (max. monitoring path length 250 meters) •
Receiver RE 130 • Optic fibre cable OF60-R (low-loss for short wavelengths) • Multiplexers MX 004 and MX 024 •
Dataloggers DL 010 and DL 016 • Analogue and digital input/output cards AO 008, AI016, and DI032 • Analogue and digital
isolation cards IA 008, ID 008, OA 008, and OD 008 • Window heaters HF 110 and HF 150 • Mirror heaters HM 110 and HM
150 • Auto calibration unit CU 007 • Software packages IO 80 (for the analogue and digital input/output adapters), DL10 and
DL16 (for data loggers), ComVision, and STAT 500;

Recommended calibration and accuracy audit components (or equivalent) for either Model AR 500 or System 300:

• Wavelength calibration lamp CA 004 • Calibration bench CB 100 • Receiver unit RE 060 (two required) • Calibration unit
CA 150, with same type lamp as used in the monitoring path emitter • Power supply PS 150 for calibration unit CA 150 •
Calibration cells CC 001-X, where X represents various cell lengths from 1 to 900 mm • Special calibration cells CC 110 or
CC 150 (for mounting directly on receiver) • Ozone generator OC 500 • Light meter LM 010.

Federal Register: Vol 60, page 21518, 05/02/1995

PCI Ozone Corporation Model LC-12 Ozone Analyzer
Automated Equivalent Method: EQOA-0382-055

"PCI Ozone Corporation Model LC-12 Ozone Analyzer," operated on a range of 0-0.5 ppm.

Federal Register: Vol. 47, page 13572, 03/31/1982

Philips PW9771 03 Analyzer

Automated Equivalent Method: EQOA-0777-023

"Philips PW9771 03 Analyzer," consisting of the following components: PW9771/00 03 Monitor with PW9724/00 Disc.-Set;
PW9750/00 Supply Cabinet; PW9750/20 Supply Unit operated on a range of 0-0.5 ppm, with or without any of the following
accessories: PW9732/00 Sampler Line Heater; PW9750/30 Frame For MTT; PW9750/41 Control Clock 60 Hz; PW9733/00
Sampler; PW9752/00 Air Sampler Manifold.

Federal Register: Vol. 42, page 38931, 08/01/1977; Vol 42, page 57156,11/01/1977

Seres Model OZ 2000 G Ozone Analyzer
Automated Equivalent Method: EQOA-0506-161

"Seres OZ 2000 G Ozone Ambient Air Analyzer," operated with a full scale range of 0 - 0.5 ppm, at any temperature in the
range of 20°C to 30°C, and with or without either of the following options: internal ozone generator, teletransmission interface.2
Federal Register: Vol 71, page 25587, 05/01/2006

SIR S.A. Model S-5014 O3 Analyzer
Automated Equivalent Method: EQOA-0207-164

"SIR S.A. Model S-5014 Photometric O3 Analyzer," operated on the 0 - 500 ppb measurement range, within an ambient
temperature range of 20°C to 30°C, with a sample inlet particulate filter, and with or without an optional PCMCIA card.
Federal Register: Vol 72, page 8985, 02/28/2007

Sutron Corporation Model 6030 O3 Analyzer
Automated Equivalent Method: EQOA-0415-222

"Sutron Model 6030 Ozone Analyzer," operated at any of the following measurement ranges: 0-0.05 ppm, 0-0.5 ppm and 0-
1.0 ppm, at any ambient temperature in the range of 5°C -45 °C, with an averaging time of 1 to 99 analyzer cycles (0 to 396
seconds), with sample flow rate of 0.5 to 1 Lpm and in accordance with the Model 6030 Ozone Analyzer Operation Manual
and with or without the following options: internal ozone generator, zero/span ports for external calibration.

Federal Register: Vol. 80, page 32114, 6/05/2015


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 33

Tanabyte Models 722, 723, 724, 725, or 726 Ambient Ozone Analyzer
Automated Equivalent Method: EQOA-0407-165

"Tanabyte Models 722,723, 724, 725, or 726 Ambient Ozone Analyzer," enclosed in either a Dual-Bay Chassis or a Single-
Bay Chassis and operated on either the 0 - 0.5 ppm or 0 -1.0 ppm measurement range, within an ambient temperature range of
20 to 30 degrees C, and with a sample inlet particulate filter installed in the sample filter holder.

Federal Register: Vol 72, page 20846, 04/26/2007

Teledyne Advanced Pollution Instrumentation, Model 400E or T400; Advanced Pollution Instrumentation, Model 400/400A; Teledyne
Monitor Labs sensor-e™ Model TML-10 Ozone Analyzers; or recordum airpointer® system module 801-004000;

Automated Equivalent Method: EQOA-0992-087

"Teledyne Advanced Pollution Instrumentation. Model 400E or T400; Advanced Pollution Instrumentation, Model 400 or 400A; or
Teledyne Monitor Labs sensor-e™ Model TML-10 Ozone Analyzef' operated on any lull scale range between 0-100 ppb1 and 0-1000
ppb, with any range mode (Single, Dual, or AutoRange), at any ambient temperature in the range of 5°C to 40°C, and with a TFE filter
or a Kynar® DFU. Models 400E, T400 and TIV1L-10: operated with a sample flow rate of 800 ± 80 cm3/min (sea level), with the
dilution factor set to 1, with Dynamic Zero ON or OFF, with Dynamic Span OFF, with Temp/Press compensation ON, and with or
without any of the following options: Internal or external sample pump, Sample/Cal valve option, Internal Zero/Span (IZS), Rack mount
with or without slides, analog input option, 4-20 mA isolated current loop output; the NumaView™ software.2 Models 400/400A:
operated with the dynamic zero and span adjustment feature (some Model 400 units only) set to OFF, and with or without any of the
following options: Zero/Span Valve option, Internal Zero/Span (IZS) option, IZS ozone generator reference feedback option, standard
serial port or Multi-drop RS-232, digital status outputs, analog outputs: 100 mV, IV, 5V, 10V, 4-20 mA current loop, optional metal
wool ozone scrubber, optional external sample pump, optional 47 mm diameter filter, optical bench heater, rack mount with slides; the
NumaView™ software, airpointer® system module 801-004000 only: operated on any full scale range between 0-0.100 ppm and 0-
1.0 ppm, with a PTFE filter element installed in the internal filter assembly, with the software setting: FRMZFEM conform mode; at any
temperature in the range of 10°C to 45°C, with either a user- or vendor-supplied vacuum pump capable of providing an absolute pressure
of 16 inches mercury or less at 2.5 sLpm; installed in the compact, thermally controlled (-40°C to + 45°C) and weather proof airpointer®
base unit with integrated data acquisition and management system mounted on a frame, pole, or wall; with or without wireless telemetry;
with or without the internal span option as module supplement consisting of ozone generator; with or without modules for other criteria
pollutants; with or without analyzer for particulate matter; with or without additional 3rd party sensors for e.g. meteorology, noise, or
traffic counting. Operated with the appropriate instrument manual. Note 2 applies to the following Teledyne Advanced Pollution Instrumentation,
Models 400E, T400, and Teledyne Monitor Labs, Inc. Sensor-e™ Model TML-10.

Federal Register: Vol. 57, page 44565, 09/28/1992; Vol 63, page 31992, 06/11/1998 Vol 67, page 57811, 09/12/2002
Latest Modification: 08/2010; 05/2013; 07/2014; 9/2015

Teledyne Advanced Pollution Instrumentation Model 265E or T265 Chemiluminescence Ozone Analyzers
Automated Equivalent Method: EQOA-0611-199 or Automated Reference Method: RFOA-0216-230

"Teledyne Advanced Pollution Instrumentation, Model 265E or T265 Chemiluminescence Ozone Analyzer," operated on any
full scale range between 0-100 ppb and 0-1000 ppb, with any range mode (Single, Dual, or AutoRange), at any operating
temperature in the range of 5°C to 40°C, and with a TFE filter or a Kynar® DFU in the sample air inlet, operated with a sample
flow rate of 500 ± 50 cm3/min (sea level), with the dilution factor set to 1, with Temp/Press compensation ON, and in
accordance with the appropriate associated instrument manual, and with or without any of the following options: Internal or
external sample pump, Sample/Cal valve option, Rack mount with or without slides, analog input option, 4-20 mA isolated
current loop output. Note 2 applies to the following Teledyne Advanced Pollution Instrumentation Models 265E and T265.
Federal Register: Vol. 76, page 62402,10/07/2011; Vol 81 page 25397, 04/28/2016
Latest Modification: 10/2012; 07/2014

Teledyne Advanced Pollution Instrumentation, Model T204 Analyzer
Automated Equivalent Method: EQOA-0514-214

"Teledyne Advanced Pollution Instrumentation, Model T204 NOX + 03 Analyzer", operated on any full scale range between
0-100 ppb and 0-500 ppb, at any operating temperature from 5°C to 40°C, with either a user-or vendor-supplied vacuum pump
capable of providing an absolute pressure of 10 inches mercury or less at 3 sLpm, in accordance with the associated instrument
manual, and with or without any of the following options: Zero/Span valves, external communication and data monitoring
interfaces. Note 2 applies to the Teledyne Advanced Pollution Instrumentation, Model T204.

Federal Register: Vol.79, pages 34734-34735, 06/18/2014
Latest Modification: 07/2014


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 34

Teledyne Advanced Pollution Instrumentation, Model 430 Ozone Analyzer
Automated Equivalent Method: EQOA-1015-229

"Teledyne Advanced Pollution Instrumentation, Model 430 Ozone Analyzer", operated with a full scale range between 0-500
ppb, at any operating temperature from 5°C to 40°C, with a sample particulate filter, with a 100-240V AC to DC power adapter
or a 12V DC source capable of providing 9 watts of power, in accordance with the associated instrument manual, and with or
without any of the following options: internal long-life pump, external long-life pump, external portable battery pack, external
communication and data monitoring interfaces.

Federal Register: Vol. 80, pages 72432,11/19/2015

Teledyne Monitor Labs/Casella/Ecotech Models ML9810/CM2010/EC9810, -11, or -12, Teledyne Monitor
Labs/Casella/Ecotech Model ML9810B/CM2010B/EC9810B, or Wedding & Associates Model 1010 Ozone Analyzers
Automated Equivalent Method: EQOA-0193-091

"Teledyne Monitor Labs, Casella Monitor, or Ecotech Models ML9810/CM2010/EC9810, ML9811/CM2011/EC9811,
ML9812/CM2012/EC9812, orML9810B/CM2010B/EC9810B or Wedding & Associates, Inc. Model 1010 Ozone Analyzers,"
operated on any full scale range between 0-0.05 ppm1 and 0-1.0 ppm, at any temperature in the range of 15°C to 35°C, with
the service switch on the secondary panel set to the In position; with the following menu choices selected: Range: 0.05ppm to
1.0ppm; Over-ranging: Enabled ox Disabled, Calibration: Manual or Timed, Diagnostic Mode: Operate; Filter Type: Kalman;
Pres/Temp/Flow Comp: On', Span Comp: Disabled, and as follows: Models ML9810/CM2010/EC9810, -11, and -12 - with
a five-micron Teflon® filter element installed internally, with the 50-pin I/O board installed on the rear panel configured at any
of the following output range settings: Voltage, 0. IV, IV, 5 V, 10V; Current, 0-20 mA, 2-20 mA, 4-20 mA; and with or without
any of the following options: Valve Assembly for External Zero/Span (EZS); Rack Mount Assembly; Internal Floppy Disk
Drive. Models ML9810B/CM2010B/EC9810B and 1010 - with either a vendor-supplied or equivalent user-supplied five
micron Teflon® filter and exhaust pump, and with or without any of the following options: Valve Assembly for External
Zero/Span (EZS); Rack Mount Assembly; 50-pin I/O board; Internal Zero/Span Assembly (IZS); hinged, fold-down front
panel. Operated with appropriate instrument manual.

Federal Register: Vol. 58, page 6964, 02/03/1993
Latest Modification: 03/2011

Thermo Electron/Thermo Environmental Instruments Models 49, 49C, 49i
Automated Equivalent Method: EQOA-0880-047

"Thermo Electron or Thermo Environmental Instruments, Inc. Model 49 U. V. Photometric Ambient O3 Analyzer," operated on a
measurement range of either 0-0.5 or 0-1.0 ppm with or without any of the following options: 49-001 Teflon Particulate Filter;
49-002 19 Inch Rack Mount; 49-100 Internal Ozone Generator for Zero, Precision, and Level 1 Span Check; 49-103 Internal
Ozone Generator for Zero, Precision, and Level 1 Span Checks With Remote Activation; 49-488 GPIB (General Purpose Interface
Bus) IEEE-488. Model 49C or 49i U.V. Photometric Ambient O3 Analyzer," operated on any measurement range between 0-
0.051 to 1.0 ppm, with any time average setting between 10 and 300 seconds, with the temperature and/or pressure compensation
on or off, within a temperature range of 20°C to 30°C, with or without any of the following options: 2 Teflon particulate filter,
Internal Zero Air Scrubber, Internal Ozonator with remote activation, Rack mounts; Model 49C: Internal Ozonator, Carrying
Handle, 4-20 mA current output, RS-232 Interface, RS-485 Interface; Model 49i: I/O expansion board, airpointer® Model 801-
004000-T, operated on any full scale range between 0-0.100 ppm and 0-1.0 ppm, with a PTFE filter element installed in the
internal filter assembly, with the software setting: FRM/FEM conform mode; at any internal temperature in the range of 10°C to
44°C, with either a user- or vendor-supplied vacuum pump capable of providing an absolute pressure of 16 inches mercury or less
at 2.5 sLpm; installed in the compact, thermally controlled (-40°C to + 45°C) and weather proof airpointer® base unit with
integrated data acquisition and management system mounted on a frame, pole, or wall; with or without wireless telemetry; with
or without the internal span option as module supplement consisting of ozone generator; with or without modules for other criteria
pollutants; with or without analyzer for particulate matter; with or without additional 3rd party sensors for e.g. meteorology, noise,
or traffic counting. Operated with the appropriate instrument manual.

Federal Register: Vol. 45, page 57168, 08/27/1980
Latest Modification: 12/2011; 10/2015


-------
CO

List of Designated Reference and Equivalent Methods. December 17.2016

Page 35

Carbon Monoxide

Beckman Model 866 CO Monitoring System
Automated Reference Method: RFCA-0876-012

"Beckman Model 866 Ambient CO Monitoring System," consisting of the following components: Pump/Sample-Handling
Module; Gas Control Panel; Model 865-17 Analyzer Unit; Automatic Zero/Span Standardize^ operated with a 0-50 ppm range,
a 13 second electronic response time, with or without any of the following options: Current Output Feature; Bench Mounting
Kit; Linearizer Circuit.

Federal Register: Vol. 41, page 36245, 08/27/1976

Bendix/Combustion Engineering Model 8501-5CA CO Analyzer
Automated Reference Method: RFCA-0276-008

"Bendix or Combustion Engineering Model 8501-5CA Infrared CO Analyzer," operated on the 0-50 ppm range and with a
time constant setting between 5 and 16 seconds, with or without any of the following options: Rack Mounting with Chassis
Slides; Rack Mounting without Chassis Slides; External Sample Pump.

Federal Register: Vol. 41, page 7450, 02/18/1976

Dasibi Model 3003 CO Analyzer
Automated Reference Method: RFCA-0381-051

"Dasibi Model 3003 Gas Filter Correlation Dasibi Environmental CO Analyzer," operated on the 0-50
sample particulate filter installed on the sample inlet line, with or without any of the following options:

3-001 Rack Mount	3-003 BCD Digital Output	3-007 Zero/Span Module

3-002 Remote Zero and Span 3-004 4-20 Milliamp Output
Federal Register: Vol. 46, page 20773, 04/07/1981

Dasibi Model 3008 CO Analyzer
Automated Reference Method: RFCA-0488-067

"Dasibi Model 3008 Gas Filter Correlation CO Analyzer," operated on the 0-50 ppm range, with a time constant setting of 60
seconds, a particulate filter installed in the analyzer sample inlet line, with or without use of the auto zero or auto zero/span
feature, and with or without any of the following options: N-0056-A RS-232-C Interface; S-0132-A Rack Mounting Slides; Z-
0176-S Rack Mounting Brackets.

Federal Register: Vol. 53, page 12073, 04/12/1988

DKK-TOA Corporation Model GFC-311E Ambient CO Analyzer
Automated Reference Method: RFCA-0907-167

"DKK-TOA Corporation Model GFC-31 IE Ambient CO Analyzer," operated with full scale fixed measurement ranges of 0-
5, 0-20, and 0-50 ppm at any environmental temperature in the range of 20°C to 30°C.

Federal Register: Vol. 72, page 56339,10/03/2007

Ecotech Serinus 30 Carbon Monoxide Analyzer or Opsis AB OPS 30 Carbon Monoxide Analyzer or Teledyne Analytical
Instruments GFC7001E Carbon Monoxide Analyzer or Tisch Environmental TE 3.0 Carbon Monoxide Analyzer
Automated Reference Method: RFCA-0509-174

"Ecotech Serinus 30 Carbon Monoxide Analyzer" or "Opsis AB OPS30 Carbon Monoxide Analyzer" or "Teledyne Analytical
Instruments GFC7001E Carbon Monoxide Analyzer" or "Tisch Environmental TE 3.0 Carbon Monoxide Analyzer," operated
in the range of 0-50 ppm, with a five-micron Teflon® filter element installed, and with the following selected: Background-
Enabled, Control Loop-Enabled, Diagnostic Mode-Operate, Pres/Temp/Flow Compensation-Enabled, Span Compensation-
Disabled, with concentration automatically corrected for temperature and pressure changes, and operated according to the
Serinus 30 Carbon Monoxide Analyzer User Manual or the OPS 30 Carbon Monoxide Analyzer User Manual or the Teledyne
Analytical Instruments GFC700 IE Carbon Monoxide Analyzer Instruction Manual, or the Tisch Environmental TE 3.0 Carbon
Monoxide Analyzer Instruction Manual as appropriate.

Federal Register: Vol. 74, page 26395, 06/02/2009
Latest Modifications: 05/2010, 05/2011, 05/2012, 08/2014

ppm range, with a
Panel


-------
CO

List of Designated Reference and Equivalent Methods. December 17.2016

Page 36

Environnement S.A. Model COllM CO Analyzer
Automated Reference Method: RFCA-0995-108

"Environnement S. A. Model COl 1M Ambient Carbon Monoxide Analyzer," operated on a Ml scale range of 0 - 50 ppm, at
any temperature in the range of 15°C to 35°C, with a 5-micron PTFE sample particulate filter, with the following software
settings: Automatic response time ON; Minimum response time set to 40 seconds (RT 13); Automatic ZERO-REF cycle
programmed every 24 hours; and with or without any of the following options:2 RS232-422 Serial Interface; Internal Printer.
Federal Register: Vol 60, page 54684,10/25/1995

Environnement S.A. Model C012M CO Analyzer
Automated Reference Method: RFCA-0206-147

"Environnement S. A Model C012M Gas Filter Correlation Carbon Monoxide Analyzer," operated with a full scale range of 0
- 50 ppm, at any temperature in the range of 10°C to 35°C, with a 5-micron PTFE sample particulate filter, with response time
ON, and with the automatic "ZERO-REF" cycle either ON or OFF.2
Federal Register: Vol 67, page 42557, 06/24/2002

Environnement S.A. Model C012e CO Analyzer
Automated Reference Method: RFCA-0915-228

"Environnement S.A. Model C012e Carbon Monoxide Analyzer", an infrared absorption spectroscopy technique operated on
a full scale range of 0 - 50 ppm, at any temperature in the range of 10°C to 35°C, with a teflon sample particulate filter with
the following software settings: Automatic response time ON; Automatic "ZERO-REF" cycle either ON or OFF and with or
without the following options: ESTEL Analog Input/Output Board, LCD color touch screen and Carbon Dioxide CO2 sensor.
Federal Register: Vol 80, page 72432,11/19/2015

Horiba Models AQM-10, AQM-11, or AOM12 CO Monitoring Systems
Automated Reference Method: RFCA-1278-033

"Horiba Models AQM-10, AQM-11, or AQM12 Ambient CO Monitoring Systems," operated on the 0-50 ppm range, with a
response time setting of 15.5 seconds, with or without any of the following options: AIC-101 Automatic Indication Corrector;
VIT-3 Non-Isolated Current Output; ISO-2 and DCS-3 Isolated Current Output.

Federal Register: Vol. 43, page 58429,12/14/1978

Horiba Model APMA-300E CO Monitoring System
Automated Reference Method: RFCA-1180-048

"Horiba Model APMA-300E Ambient Carbon Monoxide Monitoring System," operated on the 0-20 ppm1, the 0-50 ppm, or
the 0-100 ppm range with a time constant switch setting of No. 5. The monitoring system may be operated at temperatures
between 10°C and 40°C. (This method was originally designated as "Horiba Model APMA 300E/300SE Ambient Carbon
Monoxide Monitoring System.")

Federal Register: Vol. 45, page 72774,11/03/1980

Horiba Models APMA-360 or APMA-360-CE CO Monitor
Automated Reference Method: RFCA-0895-106

"Horiba Instruments Incorporated, Models APMA-360 or APMA-360-CE Ambient Carbon Monoxide Monitor," operated on
the 0-50 ppm range, with the Line Setting set to "MEASURE," with the Analog Output set to "MOMENTARY VALUE," and
with or without the following options:21) Rack Mounting Plate and Side Rails 2) RS-232 Com Port.

Federal Register: Vol. 60, page 39382, 08/02/1995

Horiba Model APMA-3 70 CO Monitor
Automated Reference Method: RFCA-0506-158

"Horiba Instruments Incorporated Model APMA-370 Ambient CO Monitor," operated with a full scale fixed measurement
range of 0 - 50 ppm, with the automatic range switching off, at any environmental temperature in the range of 20°C to 30°C.2
Federal Register: Vol 71, page 25587, 05/01/2006

MASS-CO, Model 1 CO Analyzer
Automated Reference Method: RFCA-1280-050

"MASS-CO, Model 1 Carbon Monoxide Analyzer," operated on a range of 0-50 ppm, with automatic zero and span adjustments
at time intervals not to exceed 4 hours, with or without the 100 millivolt and 5 volt output options. The method consists of the
following components: (1) Infra-2 (Uras 2) Infrared Analyzer Model 5611-200-35, (2) Automatic Calibrator Model 5869-111,


-------
CO

List of Designated Reference and Equivalent Methods. December 17.2016

Page 37

(3) Electric Gas Cooler Model 7865-222 or equivalent with pre humidifier, (4) Diaphragm Pump Model 5861-214 or equivalent,
(5) Membrane Filter Model 5862-111 or equivalent, (6) Flow Meter Model SK 1171-U or equivalent, (7) Recorder Model Mini
Comp DN 1/192 or equivalent. NOTE: This method is not now commercially available.

Federal Register: Vol. 45, page 81650,12/11/1980

Monitor Labs Model 8310 CO Analyzer
Automated Reference Method: RFCA-0979-041

"Monitor Labs Model 8310 CO Analyzer," operated on the 0-50 ppm range, with a sample inlet filter, with or without any of
the following options: 02A Zero/Span Valves 04B Pump (50 Hz)	07A Zero/Span Valve Power Supply

03 A Floor Stand	05A CO Regulator	08A Calibration Valves

04A Pump (60 Hz)	06A CO Cylinder	09A,B,C,D Input Power Transformer

Federal Register: Vol. 44, page 54545, 09/20/1979 and Vol 45, page 2700, 01/14/1980

Monitor Labs/Lear Siegler Model 8830 CO Analyzer
Automated Reference Method: RFCA-0388-066

"Monitor Labs or Lear Siegler Model 8830 CO Analyzer," operated on the 0-50 ppm range, with a five micron Teflon filter
element installed in the rear-panel filter assembly, with or without any of the following options: 2 - Zero/Span Valve Assembly;
3 - Rack Assembly; 4 - Slide Assembly; 7 - 230 VAC, 50/60 Hz.

Federal Register: Vol. 53, page 7233, 03/07/1988

MSA/LIRA Model 202S CO Analyzer System
Automated Reference Method: RFCA-0177-018

"LIRA Model 202S Air Quality Carbon Monoxide Analyzer System," consisting of a LIRA Model 202S optical bench (P/N
459839), a regenerative dryer (P/N 464084), and rack-mounted sampling system; operated on a 0-50 ppm range, with the slow
response amplifier, with or without any of the following options: Remote Meter; Remote Zero and Span Controls; 0-1, 5, 20,
Or 50 mA Output; 1-5, 4-20, Or 10-50 mA Output; 0-10 Or 100 mV Output; 0-1, 5, Or 10 Volt Output.

Federal Register: Vol. 42, page 5748, 01/31/1977

Peak Laboratories, Model 910-170 Carbon Monoxide Analyzer
Automated Equivalent Method: EQCA-0814-217

"Peak Laboratories, Model 910-170 Carbon Monoxide Analyzer", (Mercury replacement- UV photometric method) operated
on the standard range of 0-50 ppm and the lower range of 0-1 ppm, at any operating temperature from 20°C to 30°C, using a
back-flushing GC scrubber, 99.9999% nitrogen carrier gas at a gas pressure of 60-80 psig, with a column temperature of 105°C,
and a detector temperature of 265°C; inlet flow of 20-100 mL/min; in accordance with the associated instrument manual, and
with or without any of the following options: rack mount kit, internal sample pump, 4-20 mA output module, particle filter,
and data collection software.

Federal Register: Vol. 79, page 65392,11/04/2014
Latest modification: 6/2015

SIR S.A. Model S-5006 CO Analyzer
Automated Reference Method: RFCA-0708-172

"SIR S.A. Model S-5006 CO Analyzer," operated with full scale fixed measurement ranges 0-50 ppm at any environment
temperature in the range of 20°C to 30°C.

Federal Register: Vol. 73, page 40866, 07/16/2008

Teledyne Advanced Pollution Instrumentation Models 300,300E, 300EU, T300, T300U or Teledyne Monitor Labs sensor-e™ Model
TML-30 CO Analyzer; or recordum airpointer® system module 801-003000;

Automated Reference Method: RFCA-1093-093

"Teledyne Advanced Pollution Instrumentation Models 300, 300E, 300EU, T300, T300U or Teledyne Monitor Labs, Inc. sensor-e™
Model TML-30, Gas Filter Correlation Carbon Monoxide Analyzer," operated on any lull scale range between 0-10 ppm and 0-50 ppm
(0 - 0.1 ppm for Models 300EU and T300U), at any temperature in the range of 15°C to 35°C for Model 300 or 10°C to 40°C for
Models 300E, 300 EU, T300, T300U and TIV1L-30. with a 5-micron TFE filter element or a Kynar® DFU installed in the sample filter
assembly, with the dynamic zero and span adjustment set to Offfor Model 300, and with or without any of the following options:2 Option
50, Zero/Span Valves with pressurized span gas and shutoff valve; Option 51, Zero/Span Valves with pressurized span gas and shutoff
valve and Internal Zero Air Generator; Option 52, Zero/Span Valves; Option 53, Zero/Span Valves with Internal Zero Air Generator;
Rack Mount with slides; RS-232 serial port with status outputs; analog input option;; the NumaView™ software and (for Models 300E,


-------
CO

List of Designated Reference and Equivalent Methods. December 17.2016

Page 38

300EU, T300, T300U and TML-30) 4-20 mA isolated outputs, airpointer® model 801-003000 only: operated on any full scale range
betweenO-lOppm and 0-50 ppm, with a PTFE filter element installed in the internal filter assembly, with the software setting: FRM/FEM
conform mode; at any temperature in the range of 10°C to 45°C, with either a user- or vendor-supplied vacuum pump capable of providing
an absolute pressure of 16 inches mercury or less at 2.5 sLpm; installed in the compact, thermally controlled (-40°C to + 45°C) and
weather proof airpointer® base unit with integrated data acquisition and management system mounted on a frame, pole, or wall; with or
without wireless telemetry; with or without internal dilution system with internal span gas bottle; with or without modules for other
criteria pollutants; with or without analyzer for particulate matter; with or without additional 3rd party sensors for e.g. meteorology, noise,
or traffic counting. Operated with the appropriate instrument manual. Note 2 applies to the following Teledyne Advanced Pollution
Instrumentation Models 300E, 300EU, T300, T300U, and Teledyne Monitor Labs, Inc. Sensor-e™ Model TML-30.

Federal Register: Vol 58, page 58166,10/29/1993
Latest Modification: 08/2010; 05/2013; 07/2014; 9/2015

Teledyne Monitor Labs/Casella/Ecotech Model ML9830/CM2030/EC9830/EC9830T, Teledyne Monitor Labs/Casella/ Ecotech
Model ML9830B/CM2030B/EC9830B, or Wedding & Associates Model 1020 CO Analyzers
Automated Reference Method: RFCA-0992-088

"Teledyne Monitor Labs, Casella Monitor, or Ecotech Models ML9830/CM2030/EC9830 or ML9830B/CM2030B/ EC9830B, Ecotech
Model EC9830T, or Wedding & Associates, Inc. Model 1020 Carbon Monoxide Analyzer;"operated on any full scale range between 0-
5.0 ppm1 and 0-100 ppm, at any temperature in the range of 15°C to 35°C, with the service switch on the secondary panel set to the In
position, with the following menu choices selected: Range: 5.0ppm to 100.0 ppm; Over-ranging: Enabled or Disabled; Background: Not
Disabled, Calibration: Manual or Timed, Diagnostic Mode: Operate, Filter Type: Kalman; Pres/Temp/Flow Comp: On; Span Comp:
Disabled, and as follows: Model ML9830/CM2030/ EC9830/EC9830T: with a five-micron Teflon® filter element installed internally,
with the 50-pin I/O board installed on the rear panel configured at any of the following output range settings: Voltage, 0.1V, IV, 5 V,
10V; Current, 0-20 mA, 2-20 mA and 4-20 mA; and with or without any of the following options: Valve Assembly for External
Zero/Span (EZS); Valve Assembly for Internal Zero/Span (IZS); Rack Mount Assembly; Internal Floppy Disk Drive. Models
ML9830B/CM2030B/ EC9830B and 1020: with either a vendor-supplied or equivalent user-supplied five micron Teflon® filter and
exhaust pump, and with or without any of the following options: Valve Assembly for External Zero/Span (EZS); 50-pin I/O board; Rack
Mount Assembly; High Pressure Span Valve; hinged, fold-down front panel. Operated with appropriate instrument manual.

Federal Register: Vol 57, page 44565, 09/28/1992
Latest Modification: 03/2011

Thermo Electron/Thermo Environmental Instruments Models 48, 48C, 48CTLE, 48i, 48iTLE
Automated Reference Method: RFCA-0981-054

"Thermo Electron or Thermo Environmental Instruments, Inc. Model 48 Gas Filter Correlation Ambient CO Analyzer," operated on the
0-50 ppm range, with a time constant setting of 30 seconds, with or without any of the following options:

48-001 Teflon Particulate Filter	48-010 Internal Zero Air Package

48-002 19 Inch Rack Mount	48-488 GPIB (General Purpose Interface Bus) EEEE-488

48-003 Internal Zero/Span Valves with Remote Activation
"Thermo Electron or Thermo Environmental Instruments, Inc. Models 48C or 48i Gas Filter Correlation Ambient CO Analyzer,"
operated on any measurement range between 0-1 ppm1 and 0-100 ppm, with any averaging time setting fro in 10 to 300 seconds, with
temperature and/or pressure compensation on or off, operated at temperatures between 20°C and 30°C, with or without any of the
following options:2 Teflon particulate filter, Internal zero air scrubber, I/O Expansion board; Model 48C: Carrying handle, 4-20 mA
current output, Rack mounts, RS-232 interface, Internal zero/span and sample/calibration solenoid valves, RS-485 interface, Internal
zero/span and sample/calibration solenoid valves with remote I/O activation; Models 48C or 48i Trace Level-Enhanced (TLE) Gas
Filter Correlation Ambient CO Analyzers operated between 0-1 and 100 ppm with averaging time from 10 to 300 seconds, operated
at temperatures between 20° and 30°C at line voltages of 90-110,105-125, and 210-250 VAC @ 50/60 Hz, with or without any of the
following options: rack mounts, Teflon® particulate filter, I/O Expansion board, airpointer® Model 801-003000-T: operated on any
lull scale range between 0-10 ppm and 0-50 ppm, with a PTFE filter element installed in the internal filter assembly, with the software
setting: FRM/FEM conform mode; at any internal temperature in the range of 15°C to 45°C, with either a user- or vendor-supplied
vacuum pump capable of providing an absolute pressure of 16 inches mercury or less at 2.5 sLpm; installed in the compact, thermally
controlled (-40°C to + 45°C) and weather proof airpointer® base unit with integrated data acquisition and management system mounted
on a frame, pole, or wall; with or without wireless telemetry; with or without the internal span option as module supplement consisting
of an internal dilution system with internal span gas bottle; with or without modules for other criteria pollutants; with or without analyzer
for particulate matter; with or without additional 3rd party sensors for e.g. meteorology, noise, or traffic counting. Operated with the
appropriate instrument manual.

Federal Register: Vol 46, page 47002, 09/23/1981
Latest modifications: 04/2009; 10/2010; 10/2015


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 39

Nitrogen Dioxide

Sodium Arsenite Method for NO2
Manual Equivalent Method: EQN-1277-026

"Sodium Arsenite Method for the Determination of Nitrogen Dioxide in the Atmosphere."

Federal Register: Vol. 42, page 62971,12/14/1977

Sodium Arsenite Method for NO2 - Technicon II
Manual Equivalent Method: EQN-1277-027

"Sodium Arsenite Method for the Determination of Nitrogen Dioxide in the Atmosphere-Technicon II Automated Analysis System."
Federal Register: Vol. 42, page 62971,12/14/1977

TGS-ANSA Method for NO2

Manual Equivalent Method: EQN-1277-028

"TGS-ANSA Method for the Determination of Nitrogen Dioxide in the Atmosphere."

Federal Register: Vol. 42, page 62971,12/14/1977

Advanced Pollution Instrumentation, Inc. Model 200 NO2 Analyzer
Automated Reference Method: RFNA-0691-082

"Advanced Pollution Instrumentation, Inc. Model 200 Nitrogen Oxides Analyzer," operated on a range of either 0-0.5 or 0-1.0
ppm, with a 5-micron TFE filter element installed in the rear-panel filter assembly, with either a user- or vendor-supplied
vacuum pump capable of providing 5 inches mercury absolute pressure at 5 sLpm, with either a user- or vendor-supplied dry
air source capable of providing air at a dew point of 0°C or lower, with the following settings of the adjustable setup variables:

Adaptive Filter = On	PMT Temperature Set Point = 15°C	Normal Filter Size = 12 samples

Dwell Time = 7 seconds	Rate of Change (ROC) Threshold = 10%	Dynamic Span = Off

Sample Time = 8 seconds	Reaction Cell Temperature = 50°C	Dynamic Zero = Off

and with or without any of the following options:

180 Stainless Steel Valves	283 Internal Zero/Span With Valves (IZS) 356 Level One Spares Kit

184 Pump Pack	325 RS-232/Status Output	357 Level Two Spares Kit

280 Rack Mount With Slides	355 Expendables	PE5 Permeation Tube for IZS

Federal Register: Vol. 56, page 27014, 06/12/1991

Beckman Model 952-A NOZNO2/NOx Analyzer
Automated Reference Method: RFNA-0179-034

"Beckman Model 952-A NO/NCh/NOx Analyzer," operated on the 0-0.5 ppm range with the 5-micron Teflon sample filter
(Beckman P/N 861072 supplied with the analyzer) installed on the sample inlet line, with or without the Remote Operation
Option (Beckman No. 635539).

Federal Register: Vol. 44, page 7806, 02/07/1979

Bendix Model 8101-B Oxides of Nitrogen Analyzer
Automated Reference Method: RFNA-0479-038

"Bendix Model 8101-B Oxides of Nitrogen Analyzer," operated on a 0-0.5 ppm range with a Teflon sample filter installed on
the sample inlet line and with the following post-manufacture modifications: 1) Ozone generator and reaction chamber input-
output tubing modification per Bendix Service Bulletin 8101B-2; 2) The approved converter material; 3) The revised and EPA-
approved operation and service manual. These items are mandatory and must be obtained from ABB Process Analytics. The
analyzer may be operated with or without any of the following optional modifications: a. Perma Pure dryer/ambient air
modification; b. Valve cycle time modification; c. Zero potentiometer centering modification per Bendix Service Bulletin
810 IB-1; d. Reaction chamber vacuum gauge modification.

Federal Register: Vol. 44, page 26792, 05/07/1979


-------
no2

List of Designated Reference and Equivalent Methods. December 17.2016	Page 40

Bendix/Combustion Engineering Model 8101-C Oxides of Nitrogen Analyzer
Automated Reference Method: RFNA-0777-022

"Bendix or Combustion Engineering Model 8101-C Oxides of Nitrogen Analyzer," operated on a 0-0.5 ppm range with a
Teflon sample filter (Bendix P/N 007163) installed on the sample inlet line.

Federal Register: Vol. 42, page 37435, 07/21/1977

Columbia Scientific Industries Models 1600 and 5600 Analyzers
Automated Reference Method: RFNA-0977-025

"CSI Model 1600 Oxides of Nitrogen Analyzer," operated on a 0-0.5 ppm range with a Teflon sample filter (CSI P/N M951-
8023) installed on the sample inlet line, with or without any of the following options:

951-0103 Rack Ears	951-0114 Recorder Output, 5 V

951-0104 Rack Mounting Kit (Ears & Slides)	951-0115 External Pump (115 V, 60 Hz)

951-0106 Current Output, 4-20 mA (Non-Insulated) 951-8072 Molybdenum Converter Assembly (Horizontal)
951-0108 Diagnostic Output Option	951 -8074 Copper Converter Assembly (Horizontal)

951-0111 Recorder Output, 10 V	951-8079 Copper Converter Assembly (Vertical)

951-0112 Remote Zero/Span Sample Control	951-8085 Molybdenum Converter Assembly (Vertical)

NOTE: The vertical molybdenum converter assembly is standard on all new analyzers as of 1-1-87; however, use of any of the
other converter assemblies is optional. Also, the above options reflect new CSI part numbers.

"CSI Model 5600 Oxides of Nitrogen Analyzer," operated on a 0-0.5 ppm range, with any signal integration time in the range
of 20 to 99 seconds, with a Teflon sample filter (CSI P/N M951-8023) installed on the sample inlet line, and with or without
any of the following options: 954-0121 - Status Contacts; 964-0126 - Printer; 954-0131 - Rack Mounting Kit (ears and slides);
954-0122 - Input Solenoids; 954-8024 - Cartridge Dryer; 964-0012 - Single Headed Pump - Gast; 954-0125 - Current Output,
4-20 mA; 951-0115 - Single Headed Pump - KNF.

Federal Register: Vol. 42, page 46574, 09/16/1977

Dasibi Model 2108 Oxides of Nitrogen Analyzer
Automated Reference Method: RFNA-1192-089

"Dasibi Model 2108 Oxides of Nitrogen Analyzer," operated on the 0-500 ppb range, with software revision 3.6 installed in
the analyzer, with the auto thumbwheel switch and the diag thumbwheel switch settings at 0, with the following internal CPU
dipswitch settings:

switch	position	function

1	open (down)	Recorder outputs are NO & NO2

5	open (down)	3 minute time constant

6	closed (up)	3 minute time constant;

with a 5-micron Teflon filter element installed in the filter holder, and with or without any of the following options:

Built-in Permeation Oven Rack Mounting Three-Channel Recorder Output RS-232 Interface 4-20 mA Output
Federal Register: Vol. 57, page 55530,11/25/1992

DKK-TOA Corporation Model GLN-114E Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-0798-121

"DKK-TOA Corporation Models GLN-114E and GLN-114E-1 Nitrogen Oxides Analyzer," operated within a temperature
range of 20 to 30 degrees C, on any of the following measurement ranges: 0-0.0501, 0-0.1001, 0-0.2001, 0-0.500, and 0-1.000
ppm, and with or without the optional Internal zero air supply and permeation tube oven.2
Federal Register: Vol 63, page 41253, 08/03/1998

DKK-TOA Corporation Model GLN-314E Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-0508-171

"DKK-TOA Corporation Model GLN-314E Nitrogen Oxides Analyzer," operated at any temperature in the range of 20°C to
30°C, on any of the following measurement ranges:1 0-0.100 ppm, 0-0.200 ppm, 0-0.500 ppm.2
Federal Register: Vol. 73, page 28819, 05/19/2008


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 41

Ecotech Serinus 40 Oxides of Nitrogen Analyzer or Opsis AB OPS40 Oxides of Nitrogen Analyzer or Teledyne Analytical
Instruments 9110E Oxides of Nitrogen Analyzer or Tisch Environmental TE 4.0 Oxides of Nitrogen Analyzer
Automated Reference Method: RFNA-0809-186

"Ecotech Serinus 40 Oxides of Nitrogen Analyzef' or "Opsis AB OPS40 Oxides of Nitrogen Analyzef' or "Teledyne Analytical
Instruments 9110E Oxides of Nitrogen Analyzer" or "Tisch Environmental TE 4.0 Oxides of Nitrogen Analyzer," operated in
the range of 0-0.5 ppm, with a five-micron Teflon® filter element installed, with either a molybdenum or a vitreous carbon N02
to NO catalyst, and with the following selected: Control Loop-Enabled, Diagnostic Mode-Operate, Pres/Temp/Flow
Compensation-Enabled, Span Compensation-Disabled, with concentration automatically corrected for temperature and pressure
changes, and operated according to the Serinus 40 Oxides of Nitrogen Analyzer User Manual or the OPS40 Oxides of Nitrogen
Analyzer User Manual or the Teledyne Analytical Instruments 9110E Oxides of Nitrogen Analyzer Instruction Manual or Tisch
Environmental TE 4.0 Oxides of Nitrogen Analyzer Instruction Manual as appropriate.

Federal Register: Vol. 74, page 38184, 07/31/2009
Latest Modifications: 05/2010, 05/2011, 05/2012, 03/2013, 08/2014

Environnement S. A. Model AC31M NO2 Analyzer
Automated Reference Method: RFNA-0795-104

"Environnement S. A. Model AC31M Chemiluminescent Nitrogen Oxide Analyzer," operated with a full scale range of 0 - 500
ppb, at any temperature in the range of 15°C to 35°C, with a 5-micron PTFE sample particulate filter, with the following software
settings: Automatic response time ON; Minimum response time set to 60 seconds (RT) 2); and with or without any of the following
options:2 Internal Permeation Oven; Connection for Silica Gel Dryer; RS232-422 interface; EV3 valve; Internal Printer.

Federal Register: Vol 60, page 38326, 07/26/1995

Environnement S. A. Model AC32M NO2 Analyzer
Automated Reference Method: RFNA-0202-146

"Environnement S. A. Model AC32M Chemiluminescent Nitrogen Oxides Analyzer," operated with a full scale range of 0 -
500 ppb, at any temperature in the range of 10°C to 35°C, with a 5-micron PTFE sample particulate filter, with response time
setting 11 (automatic response time), and with or without the following option: Internal permeation oven.

Federal Register: Vol. 67, page 15567, 04/02/2002

Environnement S.A. SANOA Multigas Longpath Monitoring System
Automated Reference Method: EQNA-0400-139

"Environnement S.A. Model SANOA Multigas Longpath Air Quality Monitoring System," consisting of a receiver, one or
more projectors, interface unit, a user-provided control unit computer running the SANOA VisionAIR software, and associated
incidental equipment; configured for measuring NO2, with the temperature control and internal calibration cell options installed,
operated with a measurement range of 0 to 0.5 ppm, over an installed monitoring path length of between 27 and 500 meters,
within an ambient air temperature range of -30 to +45°C, with a measurement (integrating) time of 180 seconds, and with or
without external temperature and barometric pressure sensors or any of the following options: external (meteo) input
connection, series 1M bus connection, OGR type projector, analog outputs.

Federal Register: Vol. 65, page 26603, 05/08/2000

Environnement S.A. Model AS32M Nitrogen Dioxide Analyzer
Automated Equivalent Method: EQNA-1013-210

"Environnement S.A Model AS32M cavity attenuated phase shift spectroscopy Nitrogen Dioxide Analyzef', operated on any lull scale range
between 0-500 ppb and 0-1000 ppb, at any ambient temperature in the range of 20°C to 30°C, with automatic response time ON, set to 11, in
accoidance with the associated instrument manual; with sample particulate filter, zero gas inlet and zero check enabled; sample permeation diyer.
Serial link: 2 RS-232; USB port; Ethernet port (TCP/IP); onboard html web server and, with or without any of the following options: internal
permeation bench; ESTEL analog inputs/outputs.

Federal Register: Vol. 78, page 67360,11/12/2013

Horiba Instruments Models APNA-360 or APNA-360-CE NO-NO2-NOX Monitor
Automated Reference Method: RFNA-0196-111

"Horiba Instruments, Inc. Models APNA-360 or APNA-360-CE Ambient NO-NO2-NOX Monitor," operated with a full scale
range of 0 - 0.50 or 0 -1.0 ppm, at any temperature in the range of 10°C to 40°C, with a Line Setting of "MEASURE", and an
Analog Output of "MOMENTARY VALUE", and with or without the following options:2 1) Rack Mounting Plate and Side
Rails 2) RS-232 Communications Port.

Federal Register: Vol 61, page 11404, 03/20/1996


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 42

Horiba Instruments Model APNA-3 70 NO 2 Monitor
Automated Reference Method: RFNA-0506-157

"Horiba Instruments Incorporated Model APNA-370 Ambient NOx Monitor," standard specification, operated with a full scale
fixed measurement range of 0 - 0.50 ppm with the automatic range switching off, at any ambient temperature in the range of
20°C to 30°C, and with a 0.3 micrometer sample particulate filter installed.2
Federal Register: Vol 71, page 25587, 05/01/2006

Meloy Model NA530R Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-1078-031

"Meloy Model NA530R Nitrogen Oxides Analyzer," operated on the following ranges and time constant switch positions:
Range, ppm:	0-0.11	0-0.251	0-0.5	0-1.0

Time Constant Setting:	4	3 or 4	2,3, or 4	2,3, or 4

Operation of the analyzer requires an external vacuum pump, either Meloy Option N-10 or an equivalent pump capable of
maintaining a vacuum of 200 torr (22 inches mercury vacuum) or better at the pump connection at the specified sample and
ozone-air flow rates of 1200 and 200 cm3/min, respectively. The analyzer may be operated at temperatures between 10°C and
40°C and at line voltages between 105 and 130 volts, with or without any of the following options: N-l A Automatic Zero and
Span; N-2 Vacuum Gauge; N-4 Digital Panel Meter; N-6 Remote Control For Zero and Span; N-6B Remote Zero/Span Control
and Status (Pulse); N-6C Remote Zero/Span Control and Status (Timer); N-9 Manual Zero/Span; N-10 Vacuum Pump
Assembly (See Alternate Requirement Above); N-ll Auto Ranging; N-14B Line Transmitter; N-18 Rack Mount Conversion;
N-18A Rack Mount Conversion.

Federal Register: Vol. 43, page 50733,10/31/1978 and Vol 44, page 8327, 02/09/1979

Monitor Labs Model 8440E Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-0677-021

"Monitor Labs Model 8440E Nitrogen Oxides Analyzer," operated on a 0-0.5 ppm range (position 2 of range switch) with a
time constant setting of 20 seconds, with or without any of the following options:

TF-Sample Particulate Filter DO-Status Outputs 018A-Ozone Dry Air 018B-Ozone Dry Air - No Drierite
With TFE Filter Element	R- Rack Mount	V- Zero/Span Valves FM- Flow meters

Federal Register: Vol. 42, page 37434, 07/21/1977; Vol 42, page 46575, 09/16/1977; Vol 46, page 29986, 06/04/1981

Monitor Labs/Lear Siegler Model 8840 Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-0280-042

"Monitor Labs or Lear Siegler Model 8840 Nitrogen Oxides Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with
an internal time constant setting of 60 seconds, a TFE sample filter installed on the sample inlet line, with or without any of the
following options:

02 Flowmeter	08A Pump Pac Assembly with 09A (115 VAC)	011A Recorder Output 1 Volt

03A Rack Ears	08B Pump Pac Assembly with 09B (100 VAC)	01 IB Recorder Output 100 mV

03B Slides	08C Pump Pac Assembly with 09C (220/240 VAC)	011C Recorder Output 10 mV

05A Zero/Span Valves 08D Rack Mount Panel Assembly	012A DAS Output 1 Volt

05B Valve/Relay	09A Pump 115 VAC 50/60 Hz	012B DAS Output 100 mV

06 Status	09B Pump 100 VAC 50/60 Hz	012C DAS Output 10 mV

07A Input Power Transformer 100 VAC, 50/60 Hz 09C Pump 220/240 VAC 50 Hz 013A Ozone Dry Air
07B Input Power Transformer 220/240 VAC 50 Hz	013B Ozone Dry Air - No Drierite

Federal Register: Vol. 45, page 9100, 02/11/1980 and Vol 46, page 29986, 06/04/1981

Monitor Labs/Lear Siegler Model 8841 Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-0991-083

"Monitor Labs or Lear Siegler Model 8841 Nitrogen Oxides Analyzer," operated on the 0-0.05 ppm,1 0-0.1 ppm,1 0-0.2 ppm,1
0 - 0.5 ppm, or 0-1.0 ppm range, with manufacturer-supplied vacuum pump or alternative user-supplied vacuum pump capable
of providing 200 torr or better absolute vacuum while operating with the analyzer.

Federal Register: Vol. 56, page 47473, 09/19/1991


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 43

Opsis Model AR 500 and System 300 Open Path Ambient Air Monitoring Systems for NO2
Automated Equivalent Method: EQNA-0495-102

"Opsis Model AR 500 System" or "System 300 Open Path (long path) Ambient Air Monitoring Systems," configured for
measuring NO2, with one detector and movable grating, operated with a measurement range of 0 to 0.5 ppm, an installed
monitoring path length between 50 and 500 meters (or 50 and 1000 meters with the ER 150 option, AR 500 System only),
xenon lamp type B (150 watt), fiber optic cable length between 3 and 20 meters; operating within an ambient air temperature
range of -50 to +50°C, an analyzer temperature range of 20 to 30°C, a measurement (integrating) time setting between 30 and
120 seconds (0 min:30 sec. to 2 min:00 sec.), and with a complete cycle time of not more than 200 seconds (3 min, 20 sec.).
Under this method designation, the Model AR 500 System or System 300 consists of: AR 500 opto-analyser; emitter EM 110
and receiver RE 110 (together identified as ER 110); optic fibre cable OF60-S; power supply PS 150; Opsis operational
software, version 7.0 or 7.1; and initial on-site installation, setup, and limited operator training.2

Optional components that can be used with the Model AR 500 only, in addition to or as alternative to corresponding
components listed above: • AR 503 opto-analyzer configured as Model AR 500 (only the center detector active, sequential
monitoring) • Emitter/receiver ER 150 (for monitoring path lengths up to 1 kilometer) • Transceiver ER 130 and Retroreflector
RE 090 with 7 prisms (max. monitoring path length 150 meters) or 12 prisms (max. monitoring path length 250 meters) • Receiver
RE 130 • Xenon lamp type A (higher short-wavelength UV output) • Optic fibre cable OF60-R (low-loss for short wavelengths)
• Multiplexers MX 004 and MX 024 • Dataloggers DL 010 and DL 016 • Analogue and digital input/output cards AO 008, AI
016, and DI032 • Analogue and digital isolation cards IA 008, ID 008, OA 008, and OD 008 • Window heaters HF 110 and HF
150 • Mirror heaters HM 110 and HM 150 • Auto calibration unit CU 007 • Software packages IO 80 (for the analogue and
digital input/output adapters), DL10 and DL16 (for data loggers), ComVision, and STAT 500.

Recommended calibration and accuracy audit components (or equivalent) for either Model AR 500 or System 300: •

Wavelength calibration lamp CA 004 • Calibration bench CB 100 • Receiver unit RE 060 (two required) • Calibration unit CA
150, with same type lamp as used in the monitoring path emitter • Power supply PS 150 for calibration unit CA 150 • Calibration
cells CC 001-X, where X represents various cell lengths from 1 to 900 mm • Filter GG 400 • Special calibration cells CC 110
or CC 150 (for mounting directly on receiver) • Light meter LM 010
Federal Register: Vol 60, page 21518, 05/02/1995

Philips Model PW9762/02 NO/NO:/N(h Analyzer
Automated Reference Method: RFNA-0879-040

"Philips Model PW9762/02 NO/NO/NOx Analyzer," consisting of the following components: PW9762/02 Basic Analyzer;
PW9729/00 Converter Cartridge; PW9731/00 Sampler or PW9731/20 Dust Filter; operated on a range of 0-0.5 ppm, with or
without any of the following accessories: PW9752/00 Air Sampler Manifold; PW9732/00 Sample Line Heater; PW9011/00
Remote Control Set.

Federal Register: Vol. 44, page 51683, 09/04/1979

Seres Model NOx 2000 G Nitrogen Dioxide Analyzer
Automated Reference Method: RFNA-0706-163

"Seres Model NOx 2000 G Nitrogen Dioxide Ambient Air Analyzer," operated with a full scale measurement range of 1 - 0.50
ppm, at any ambient temperature in the range of 20°C to 30°C.

Federal Register: Vol. 71, page 42089, 07/25/2006

SIR S.A. Model S-5012 Nitrogen Oxides Analyzer
Automated Reference Method: RFNA-0804-152

"SIR S.A. Model S-5012 Chemiluminescent Nitrogen Oxides Analyzer," operated with a full scale range of 0 - 500 ppb, at any
temperature in the range of 20°C to 30°C, with the integration time set to 1 minute, with the "initial zero" disabled, and with a
specified Teflon particulate filter installed in the sample inlet filter holder.2
Federal Register: Vol. 69, page 47924, 08/06/2004

Teledyne Advanced Pollution Instrumentation Models 200A, 200'AU, 200E, 200EU, T200, T200U, T204; Teledyne
Analytical Instruments Model 911 OA; or Teledyne Monitor Labs sensor-e™ Model TML-41 NO2 Analyzers; or recordum
airpointer® System, Module 801-002000;

Automated Reference Method: RFNA-1194-099

"Teledyne Advanced Pollution Instrumentation Models 200A, 200AU, 9110A, 200E, 200EU, T200, T200U, T204; Teledyne
Analytical Instruments Model 9110A; or Teledyne Monitor Labs, Inc. sensor-e™ Model TML-41 Chemiluminescence Nitrogen
Oxides Analyzer," operated on any full scale range between 0-0.05 ppm and 0-1.0 ppm, with a PTFE filter element or a Kynar® DFU
installed in the internal filter assembly, with the following software settings: dynamic zero: OFF or ON; dynamic span: OFF; cal-on-


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 44

NO2: OFF; dilution factor: OFF or set to 1.0; autocal: ON or OFF; independent range: ON or OFF; autorange: ON or OFF;
temperature/pressure compensation: ON; and with or without any of the following options (if available): rack mounts with or without
slides, rack mount for external pump, zero/span valves, 4-20 mA analog outputs, status outputs, RS-232 output; the NumaView™
software. Models 200A, 200E, and T200 and TML-41 only: operated at any temperature in the range of 5°C to 40°C, with either a
user- or vendor-supplied vacuum pump capable of providing an absolute pressure of 10 inches mercury or less at 2 sLpm, with or
without optional internal zero/span (IZS) and permeation tubes for IZS, gold-plated reaction chamber, or Nafion-type sample gas
conditioner, ethernet output, control input, analog input option, RS-485 output. Model 200AU, 200EU, and T200U only: operated at
any temperature in the range of 20°C to 30°C, with either a user- or vendor-supplied vacuum pump capable of providing an absolute
pressure of 4 inches mercury or less at 1 sLpm Model T204 NOX + 03 Analyzer only: operated on any full scale range between 0-
100 ppb and 0-500 ppb, at any operating temperature from 5°C to 40°C, with either a user-or vendor-supplied vacuum pump capable
of providing an absolute pressure of 10 inches mercury or less at 3 sLpm, in accordance with the associated instrument manual, and
with or without any of the following options: Zero/Span valves, external communication and data monitoring interfaces. Airpointer®
system module 801-002000 only: operated on any full scale range between 0-0.05 ppm and 0-1.0 ppm, with a PTFE filter element
installed in the internal filter assembly, with the software setting: FRM/FEM conform mode; at any temperature in the range of 10°C
to 45°C, with either a user- or vendor-supplied vacuum pump capable of providing an absolute pressure of 16 inches mercury or less at
2.5 sLpm; installed in the compact, thermally controlled (-40°C to + 45°C) and weather proof airpointer® base unit with integrated data
acquisition and management system mounted on a frame, pole, or wall; with or without wireless telemetry; with or without internal
span option as module supplement consisting of permeation oven and permeation tube; with or without modules for other criteria
pollutants; with or without analyzer for particulate matter; with or without additional 3rd party sensors for e.g. meteorology, noise, or
traffic counting. Operated with the appropriate instrument manual. Note 2 applies to the following Teledyne Advanced Pollution
Instrumentation Models 200E, 200EU, T200, T200U, T204 and Teledyne Monitor Labs, Inc. Sensor-e™ Model TML41.

Federal Register: Vol. 59, page 61892,12/02/1994

Latest modifications: 03/2009; 08/2010; 10/2012; 5/2013; 06/2014; 07/2014; 9/2015

Teledyne Advanced Pollution Instrumentation Model 200EUP or T200UP Chemiluminescence Nitrogen Oxides Analyzer
Automated Equivalent Method: EQNA-0512-200

"Teledyne Advanced Pollution Instrumentation Model 200EUP or T200UP Chemiluminescence Nitrogen Oxides Analyzer,"
operated on any full scale range between 0-50 ppb and 0-1000 ppb, with any range mode (Single, Independent, or AutoRange),
at any ambient temperature in the range of 20°C to 30°C, equipped with either the nominal or high efficiency photolytic
converter, with software Temperature and Pressure compensation ON, in accordance with the associated instrument manual;
and with or without any of the following options: Zero/Span Valves, standard serial port (RS232/RS485) or Multi-drop RS-
232, Ethernet port, USB COM port, analog inputs, digital status outputs, analog outputs: 100 mV, IV, 5V, 10V, 4-20 mA
current loop outputs; with the Numaview™ software,. Note 2 applies to the following Teledyne Advanced Pollution Instrumentation,
Models 200EUP and T200UP.

Federal Register: Vol 77, page 32632, 06/01/2012
Latest Modification: 10/2012; 07/2014; 5/2016; 11/2016

Teledyne Advanced Pollution Instrumentation, Model T500U CAPS Nitrogen Dioxide Analyzer
Automated Equivalent Method: EQNA-0514-212

"Teledyne Advanced Pollution Instrumentation, Model T500U cavity attenuated phase shift spectroscopy Nitrogen Dioxide Analyzef',
operated on any full scale range between 0-50 ppb and 0-1000 ppb, with any range mode (Single, Dual, or AutoRange), at any operating
temperature from 5°C to 40°C, with a sample particulate filter, with the following software setting: Temperature and Pressure
compensation ON; in accordance with the associated instrument manual, and with or without any of the following options: Zero/Span
valves, internal Zero/Span permeation oven (IZS), external communication and data monitoring interfaces; and the NumaView™
software. Note 2 applies to the Teledyne Advanced Pollution Instrumentation, Model T500U.

Federal Register: Vol.79, pages 34734-34735, 06/18/2014
Latest modification: 11/2015

Teledyne Advanced Pollution Instrumentation Model T200P Chemiluminescence Nitrogen Oxides Analyzer
Automated Equivalent Method: EQNA-1016-241

"Teledyne Advanced Pollution Instrumentation Model T200P Chemiluminescence Nitrogen Oxides Analyzer," operated on
any full scale range between 0-50 ppb and 0-1000 ppb, with a PTFE filter element or a Kynar® DFU installed in the filter
assembly, with any range mode (Single or Dual), at any operating temperature in the range of 15°C to 35°C, with the high
efficiency photolytic converter, with software Temperature and Pressure compensation ON, in accordance with the associated
instrument manual; and with or without any of the following options: Zero/Span valves, internal Zero/Span permeation oven
(IZS), Nafion-type sample gas conditioner, external communication and data monitoring interfaces; and the NumaView™
software. Note 2 applies to the Teledyne Advanced Pollution Instrumentation, Model T200P.

Federal Register: Vol.81, pages 85561-85562,11/28/2016


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 45

Teledyne Monitor Labs/Casella/Ecotech Models ML9841/CM2041, ML9841A/CM2041A/EC9841A/EC9841T, Teledyne
Monitor Labs/Casella/Ecotech ModelML9841B/CM2041B/EC9841B, or Wedding & Associates Model 1030 NO2 Analyzers
Automated Reference Method: RFNA-1292-090

"Teledyne Monitor Labs, Casella Monitor, or Ecotech Models ML9841/CM2041, ML9841A/CM2041A/EC9841A, or
ML9841B/CM2041B/EC9841B, Ecotech Model 984IT, or Wedding & Associates, Inc. Model 1030 Nitrogen Oxides
Analyzers," with either a molybdenum or a vitreous carbon N02 to NO catalyst, operated on any full scale range between 0-
0.05 ppm1 and 0-1.0 ppm, at any temperature in the range of 15°C to 35°C, with the service switch on the secondary panel set
to the In position; with the following menu choices selected: Range: 0.05 ppm to 1.0ppm; Over-ranging: Enabled or Disabled,
Calibration: Manual or Timed, Diagnostic Mode: Operate; Filter Type: Kalman; Pres/Temp/Flow Comp: On; Span Comp:
Disabled, and as follows: Models ML9841/CM2041, ML9841A/CM2041A/EC9841A, and EC9841T - with a five-micron
Teflon® filter element installed internally, with the 50-pin I/O board installed on the rear panel configured at any of the
following output range setting: Voltage, 0.1V, IV, 5V, 10V; Current, 0-20 mA, 2-20 mA, 4-20 mA; and with or without any
of the following options: Valve Assembly for External Zero/Span (EZS); Internal Zero/Span (IZS) Assembly for; Rack Mount
Assembly; Internal Floppy Disk Drive. Models ML9841B/CM2041B/EC9841B and 1030 - with a vendor-supplied or
equivalent user-supplied five-micron Teflon® filter and exhaust pump, and with or without any of the following options: Valve
Assembly for External Zero/Span (EZS); 50-pin I/O board; Internal Zero/Span (IZS) Assembly; Rack Mount Assembly;
Charcoal exhaust scrubber; hinged, fold-down front panel. Operated with appropriate instrument manual.

Federal Register: Vol. 57, page 60198,12/18/1992
Latest Modification: 03/2011, 03/2013

Thermo Electron/Thermo Environmental Instruments Model 14 B/E
Automated Reference Method: RFNA-0179-035

"Thermo Electron or Thermo Environmental Instruments, Inc. Model 14 B/E Chemiluminescent NO/NO2/NOX Analyzer,"
operated on the 0-0.5 ppm range, with or without any of the following options:

14-001 Teflon Particulate Filter 14-003 Long-Time Signal Integrator	14-005 Sample Flowmeter

14-002 Voltage Divider Card 14-004 Indicating Temperature Controller	14-006 Air Filter

Federal Register: Vol. 44, page 7805, 02/07/1979 and Vol.44, page 54545, 09/20/1979

Thermo Electron/Thermo Environmental Instruments Model 141)/E
Automated Reference Method: RFNA-0279-037

"Thermo Electron or Thermo Environmental Instruments, Inc. Model 14 D/E Chemiluminescent NO/NO2/NOX Analyzer," operated on
the 0-0.5 ppm range, with or without any of the following options: 14-001 Teflon Particulate Filter, 14-002 Voltage Divider Card.
Federal Register: Vol. 44, page 10429, 02/20/1979

Thermo Environmental Instruments Models 42, 42C, 42i NO/N02/NOx Analyzer
Automated Reference Method: RFNA-1289-074

"Thermo Environmental Instruments Inc. Model 42, Model 42C, or Model 42i Chemiluminescence NO-N02-NOx Analyzer,"
operated on any measurement range between 0-50 ppb and 0-1000 ppb, with any time average setting from 10 to 300 seconds,
with temperature and/or pressure compensation on or off, operated at temperatures between 15°C and 35°C, and with or without
an exhaust ozone scrubber or any of the following options: 2 Rack mounts; Internal Zero/span and sample valves with remote
activation; Ozone particulate filter; Teflon particulate filter; Ozone permeation dryer; Permeation Oven; RS-232/485 interface,
4-20 mA current output, or I/O expansion board; Model 42 only: Pressure transducer, Sample/ozone flow meters. Model 42/'-
TL operated between 10 and 1000 ppb with averaging times from 10 to 300 seconds, operated at temperatures between 15°C
and 35°C at line voltages of:

a)	90-110 VAC @ 50/60/Hz

b)	105-125 VAC @ 50/60/Hz

c)	210-250 VAC @ 50/60/Hz

and with or without the following options: rack mounts, Teflon® Particle Filter, I/O Expansion Board, airpointer® Model
801-002000-T, operated on any full scale range between 0-0.05 ppm and 0-1.0 ppm, with a PTFE filter element installed in the
internal filter assembly, with the software setting: FRM/FEM conform mode; at any internal temperature in the range of 15°C
to 39°C, with either a user- or vendor-supplied vacuum pump capable of providing an absolute pressure of 16 inches mercury
or less at 2.5 sLpm; installed in the compact, thermally controlled (-40°C to + 45°C) and weather proof airpointer® base unit
with integrated data acquisition and management system mounted on a frame, pole, or wall; with or without wireless telemetry;
with or without internal span option as module supplement consisting of permeation oven and permeation tube; with or without


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 46

modules for other criteria pollutants; with or without analyzer for particulate matter; with or without additional 3rd party sensors
for e.g. meteorology, noise, or traffic counting. Operated with the appropriate instrument manual.

Federal Register: Vol. 54, page 50820,12/11/1989
Latest modification: 01/2010; 10/2015


-------
Pb

List of Designated Reference and Equivalent Methods. December 17.2016

Page 47

Lead

Inductively Coupled Plasma Mass Spectroscopy (ICP-MS)

Manual Reference Method: 40 CFRPart 50, Appendix G: RFLA-0813-813

Reference Method for the Determination of Lead in Suspended Particulate Matter Collected from Ambient Air.

Federal Register: Vol. 78, page 40000, 7/03/2013

Energy-Dispersive X-Ray Fluorescence Spectrometry (TNRCC)

Manual Equivalent Method: EQL-0783-058

"Determination of Lead Concentration in Ambient Particulate Matter by Energy-Dispersive X-Ray Fluorescence Spectrometry
(Texas Natural Resource Conservation Commission)," Texas Natural Resource Conservation Commission, P.O. Box 13087,
Austin, TX 78711-3087.

Federal Register: Vol. 48, page 29742, 06/28/1983

Energy-Dispersive X-Ray Fluorescence Spectrometry (NEA, Inc.)

Manual Equivalent Method: EQL-0589-072

"Determination of Lead Concentration in Ambient Particulate Matter by Energy-Dispersive X-Ray Fluorescence Spectrometry
(NEA, Inc.)," Nuclear Environmental Analysis, Inc., Suite 260, 10950 SW 5th Street, Beaverton, OR 97005.

Federal Register: Vol. 54, page 20193, 05/10/1989

Flame Atomic Absorption Spectroscopy (FAAS)

Manual Equivalent Method: EQLA-0813-803

Equivalent Method for the Determination of Lead in Suspended Particulate Matter Collected from Ambient Air.

Federal Register: Vol. 78, page 40000, 7/03/2013

Flame Atomic Absorption Spectrometry
Manual Equivalent Method: EQL-0380-043

"Determination of Lead Concentration in Ambient Particulate Matter by Flame Atomic Absorption Spectrometry Following
Ultrasonic Extraction with Heated HNO3-HCI"

Federal Register: Vol. 45, page 14648, 03/06/1980

Flameless Atomic Absorption Spectrometry (EPA/RTP, N.C.)

Manual Equivalent Method: EQL-0380-044

"Determination of Lead Concentration in Ambient Particulate Matter by Flameless Atomic Absorption Spectrometry
(EPA/RTP, NC)"

Federal Register: Vol. 45, page 14648, 03/06/1980

Flameless (Graphite Furnace) Atomic Absorption (Houston, Texas)

Manual Equivalent Method: EQL-0895-107

"Determination of Lead Concentration in Ambient Particulate Matter by Flameless (Graphite Furnace) Atomic Absorption
(City of Houston, Texas)," Health and Human Services Department, Environmental Chemistry Service, 1115 S. Braeswood,
Houston, TX 77030.

Federal Register: Vol 60, page 39383, 08/02/1995

Flameless Atomic Absorption Spectrometry (Omaha)

Manual Equivalent Method: EQL-0785-059

"Determination of Lead Concentration in Ambient Particulate Matter by Flameless Atomic Absorption Spectrometry (Omaha-
Douglas County Health Department)," Omaha-Douglas County Health Department, 1819 Farnam Street, Omaha, NE 68183.
Federal Register: Vol 50, page 37909, 09/18/1985

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Doe Run)

Manual Equivalent Method: EQL-0196-113

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Doe Run Co.)," Doe Run Company, Smelting Division, 881 Main Street Herculaneum, MO 63048.

Federal Register: Vol. 61, page 11404, 03/20/1996


-------
Pb

List of Designated Reference and Equivalent Methods. December 17.2016

Page 48

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (EPA/RTP)

Manual Equivalent Method: EQL-0380-045

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission

Spectrometry (EPA/RTP, NC)"

Federal Register: Vol. 45, page 14648, 03/06/1980

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (IL)

Manual Equivalent Method: EQL-1193-094

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Illinois)," State of Illinois, Environmental Protection Agency, Champaign Inorganic Laboratory, 2120
South First Street, Champaign, IL 61820
Federal Register: Vol. 58, page 61902,11/23/1993

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Kansas)

Manual Equivalent Method: EQL-0592-085

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Kansas)," State of Kansas, Department of Health and Environment, Forbes Field, Building 740, Topeka,
KS 66620-0001.

Federal Register: Vol. 57, page 20823, 05/15/1992

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Montana)

Manual Equivalent Method: EQL-0483-057

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Montana)," State of Montana, Department of Health and Environmental Sciences, Cogswell Building,
Helena, MT 59620.

Federal Register: Vol. 48, page 14748, 04/05/1983

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (NET!)

Manual Equivalent Method: EQL-1188-069

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Northern Engineering and Testing, Inc.)," Northern Engineering and Testing, Inc., P.O. Box 30615, Billings,
MT 59107.

Federal Register: Vol. 53, page 44947,11/07/1988

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (NH)

Manual Equivalent Method: EQL-1290-080

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of New Hampshire)," State of New Hampshire, Department of Environmental Services, Laboratory
Service Unit, 6 Hazen Drive (P.O. Box 95), Concord, NH 03302-0095.

Federal Register: Vol. 55, page 49119,11/26/1990

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (PA)

Manual Equivalent Method: EQL-0592-086

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Commonwealth of Pennsylvania)," Commonwealth of Pennsylvania, Department of Environmental Resources,
P.O. Box 2357, Harrisburg, PA 17105-2357.

Federal Register: Vol. 57, page 20823, 05/15/1992

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Pima, AZ)

Manual Equivalent Method: EQL-0995-109

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Pima County, Arizona)," Pima County, Wastewater Management Department, 201 North Stone Avenue,
Tucson, Arizona 85701-1207.

Federal Register: Vol 60, page 54684,10/25/1995


-------
Pb

List of Designated Reference and Equivalent Methods. December 17.2016

Page 49

Inductively Coupled Argon Plasma-Mass Spectrometry (Pima Co., AZ)

Manual Equivalent Method: EQL-0995-110

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Plasma-Mass Spectrometry
(Pima County, Arizona)," Pima County, Wastewater Management Department, 201 North Stone Avenue, Tucson, Arizona
85701-1207.

Federal Register: Vol 60, page 54684,10/25/1995

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (RI)

Manual Equivalent Method: EQL-0888-068

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Rhode Island)," State ofRhode Island Department of Health, Air Pollution Laboratory, 50 Orms Street,
Providence, RI 02904

Federal Register: Vol. 53, page 30866, 08/16/1988

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Silver Valley)

Manual Equivalent Method: EQL-1288-070

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Silver Valley Laboratories)," Silver Valley Laboratories, Inc., P.O. Box 929, Kellogg, ID 83837.

Federal Register: Vol. 53, page 48974,12/05/1988

Inductively Coupled Argon Plasma-Atomic Emission Spectrometry (TNRCC)

Manual Equivalent Method: EQL-0400-140

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Atomic Emission
Spectrometry (TNRCC)," Texas Natural Resource Conservation Commission Laboratory, 5144 E. Sam Houston Parkway N.,
Houston, TX 77030.

Federal Register: Vol. 65, page 26603, 05/8/2000

Inductively Coupled Argon Plasma-Optical Emission Spectrometry (WV)

Manual Equivalent Method: EQL-0694-096

"Determination of Lead Concentration in Ambient Particulate Matter by Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of West Virginia)," State of West Virginia, Department of Commerce, Labor and Environmental
Resources, Division of Environmental Protection, 1558 Washington Street East, Charleston, WV 25311-2599
Federal Register: Vol. 59, page 29429, 06/07/1994

Wavelength Dispersive X-Ray Fluorescence Spectrometry (CA)

Manual Equivalent Method: EQL-0581-052

"Determination of Lead Concentration in Ambient Particulate Matter by Wavelength Dispersive X-Ray Fluorescence
Spectrometry," California Department of Health Services, Air & Industrial Hygiene Laboratory, 2151 Berkeley Way, Berkeley,
CA 94704.

Federal Register: Vol. 46, page 29986, 06/04/1981

Inductively Coupled Plasma - Mass Spectrometry (Inter-Mountain Labs, Inc.)

Manual Equivalent Method: EQL-0310-189

"Procedure for Determination of Lead in Ambient TSP by Hot Plate Acid Extraction and ICP-MS Analysis," where total
suspended particulate matter (TSP ) is collected according to 40 CFR Appendix B to part 50, EPA Reference Method for the
Determination of Suspended Particulate Matter in the Atmosphere (High-Volume Method), extracted on a hot plate with 3M
HNO3 according to 40 CFR Appendix G to part 50, EPA Reference Method for the Determination of Lead in Suspended
Particulate Matter Collected from Ambient Air, and analyzed by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
based on EPA SW-846 Method 6020A.

Federal Register: Vol. 75, page 9894, 03/04/2010


-------
Pb

List of Designated Reference and Equivalent Methods. December 17.2016

Page 50

Inductively Coupled Plasma - Mass Spectrometry (US EPA/OAQPS)

Manual Equivalent Method: EQL-0510-191

"Determination of Lead Concentration in TSP by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) with Heated
Ultrasonic Nitric and Hydrochloric Acid Filter Extraction," where total suspended particulate matter (TSP) is collected
according to 40 CFR Appendix B to part 50, EPA Reference Method for the Determination of Suspended Particulate Matter in
the Atmosphere (High-Volume Method), extracted with a solution of nitric and hydrochloric acids, heated to 80° C and sonicated
for one hour, brought to a final volume of 40mL, and analyzed by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
based on EPA SW-846 Method 6020A.

Federal Register: Vol.75, page 30022, 05/28/2010

Inductively Coupled Plasma - Mass Spectrometry (US EPA/Region 9)

Manual Equivalent Method: EQL-0710-192

"Heated Nitric Acid Hot Block Digestion and ICP/MS Analysis for Lead (Pb) on TSP High-Volume Filters." In this method,
total suspended particulate matter (TSP) is collected on glass fiber filters according to 40 CFR Appendix B to part 50, EPA
Reference Method for the Determination of Suspended Particulate Matter in the Atmosphere (High-Volume Method), extracted
with a solution of nitric acid, heated on a hot block to 95°C for one hour, and brought to a final volume of 50 mL. The lead
content of the sample extract is analyzed by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS) based on EPA Method
200.8 and SW-846 Method 6020A.

Federal Register: Vol.75, page 45627, 08/03/2010

Inductively Coupled Plasma-Atomic Emission Spectroscopy (US EPA/OAQPS)

Manual Equivalent Method: EQL-0311-196

"Heated Ultrasonic Nitric and Hydrochloric Acid Digestion and ICP/AES Analysis for Lead (Pb) on TSP High-Volume
Filters." A sample of total suspended particulate matter (TSP) is collected on a glass fiber filter, using the sampler and procedure
of the EPA Reference Method for the Determination of Suspended Particulate Matter in the Atmosphere (High-Volume
Method) (40 CFR 50, Appendix B). The TSP sample is extracted with a solution of nitric and hydrochloric acid, heated in an
ultrasonic bath to 80°C for one hour, and brought to a final volume of 40 mL. The lead content of the sample extract is analyzed
by Inductively Coupled Plasma-Atomic Emission Spectrometry (ICP-AES), based on EPA SW-846 Method 6010C.

Federal Register: Vol 76, page 15974, 03/22/2011

Inductively Coupled Plasma- Mass Spectrometry (Eastern Research Group, Inc.)

Manual Equivalent Method: EQL-0512-201

"Determination of Lead in TSP by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) with Hot Block Dilute Acid and
Hydrogen Peroxide Filter Extraction." In this method, total suspended particulate matter (TSP) is collected on glass fiber filters
according to 40 CFR Appendix G to part 50, EPA Reference Method for the Determination of Lead in Suspended Particulate
Matter Collected From Ambient Air. The filter samples are extracted in a hot block at 95°C with a solution of dilute
hydrochloric acid and nitric acid and two aliquots of hydrogen peroxide, for a total of two and a half hours extraction time. The
samples are brought to a final volume of 50 mL and the lead content of the sample extract is analyzed by Inductively Coupled
Plasma-Mass Spectrometry (ICP-MS) based on EPA Compendium Method 10-3.5 and SW-846 Method 6020A.

Federal Register: Vol 77, page 32632, 06/01/2012

Inductively Coupled Plasma- Mass Spectrometry (Eastern Research Group, Inc.)

Manual Equivalent Method: EQL-0512-202

"Determination of Lead in PMio by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) with Hot Block Dilute Acid
and Hydrogen Peroxide Filter Extraction." In this method, PMio particulate matter is collected on Teflon® membrane filters
according to 40 CFR Appendix Q to part 50, EPA Reference Method for the Determination of Lead in Particulate Matter as
PMio Collected From Ambient Air. The filter samples are extracted in a hot block at 95°C with a solution of hydrochloric acid,
nitric acid, and hydrofluoric acid and an aliquot of hydrogen peroxide for a total of two and a half hours extraction time.
Samples are brought to a final volume of 50 mL and analyzed by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
based on EPA Compendium Method 10-3.5 and SW-846 Method 6020A.

Federal Register: Vol 77, page 32632, 06/01/2012


-------
Pb

List of Designated Reference and Equivalent Methods. December 17.2016

Page 51

Inductively Coupled Plasma- Mass Spectrometry (South Coast Air Quality Management District Laboratory)

Manual Equivalent Method: EQL-0514-213

"Determination of Lead (PB) on TSP Hi-Volume Filters by Microwave Assisted Digestion and Time of Flight Inductively
Coupled Plasma Mass Spectrometry (TOFICP-MS)." A sample of total suspended particulate matter (TSP) is collected on a
glass fiber filter, using the sampler and procedure of the EPA Reference Method for the Determination of Suspended Particulate
Matter in the Atmosphere (High-Volume Method) (40 CFR 50, Appendix B). The TSP sample is extracted with a solution of
nitric and digested in a microwave digestion system at 175° C for 15 minutes, centrifuged for 30 minutes at 2000 rpm, and
brought to a final acid concentration of 2-3%. The lead content of the sample extract is analyzed by Time of Flight Inductively
Coupled Plasma Mass Spectrometry (TOF ICP-MS), based on EPA SW-846 Method 6010C.

Federal Register: Vol.79, pages 34734-34735 06/18/2014


-------
Notes

List of Designated Reference and Equivalent Methods. December 17.2016

Page 52

Notes

1	Users should be aware that designation of this analyzer for operation on ranges less than the range specified in the performance
specifications for this analyzer (40 CFR 53, Subpart B) is based on meeting the same absolute performance specifications
required for the specified range. Thus, designation of these lower ranges does not imply commensurably better performance
than that obtained on the specified range.

2	This analyzer is approved for use, with proper factory configuration (if applicable), on either 50 or 60 Hertz line frequency
and nominal power line voltages of 115 VAC and 230 VAC, or similar voltages as specified in the operation or instruction
manual associated with the method.

a For cases with two designation numbers, these methods are dual designated as FRM and FEM. However, where a method
is dual designated, EPA considers the method an FRM for purposes of ensuring QA requirements are met.
b Method code as available in the AQS data system


-------
List of Designated Reference and Equivalent Methods. December 17.2016	Page 53

U.S. EPA Reference and Equivalent Methods for Ambient Air

Particulate Matter - TSP

TSP Manual Method

Method

Designation No.

Method Code"

Reference method (high-volume)



802


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 54

Particulate Matter - PM10

PM10 Samplers

Method

Designation No.

Method Code"

Andersen Instruments RAAS10-100

RFPS-0699-130

130

Andersen Instruments RAAS10-200

RFPS-0699-131

131

Andersen Instruments RAAS10-300

RFPS-0699-132

132

BGI or Mesa Laboratories Incorporated Model PQ100

RFPS-1298-124

124

BGI or Mesa Laboratories Incorporated Model PQ200

RFPS-1298-125

125

Ecotech Model 3000 PM10 High Volume Sampler

RFPS-0706-162

162

Met One Instruments, Inc. E-FRM- PM10

RFPS-0216-231

231

New Star Environmental Model NS-6070

RFPS-0202-141

141

Oregon DEQ Medium volume sampler

RFPS-0389-071

071

Rupprecht & Patashnick Partisol® 2000

RFPS-0694-098

098

R & P Partisol®-FRM 2000

RFPS-1298-126

126

R & P Partisol®-Plus 2025 Seq.

RFPS-1298-127

127

Sierra-Andersen/GMW 1200

RFPS-1287-063

063

Sierra-Andersen/GMW 321-B

RFPS-1287-064

064

Sierra-Andersen/GMW 321-C

RFPS-1287-065

065

Sierra-Andersen/GMW SA241, SA241M Dichot.

RFPS-0789-073

073

Thermo Scient

fic Partisol® 2000

RFPS-0694-098

098

Thermo Scient

fic Partisol® 2000-D, Partisol® 2000/-D

EQPS-311-197

197

Thermo Scient

fic Partisol® 2000-FRM, Partisol® 2000/

RFPS-1298-126

126

Thermo Scient

fic Partisol®-Plus 2025 Seq., Partisol® 2025/ Seq

RFPS-1298-127

127

Thermo Scient

fic Partisol®-Plus 2025-D Seq., Partisol® 2025/-D Seq

EQPS-0311-198

198

Tisch Environmental Model TE-6070

RFPS-0202-141

141

Tisch Environmental Model TE-Wilbur 10

RFPS-0714-216

216

W&A/Thermo Electron Model 600 HVL

RFPS-1087-062

062



PM10 Analyzers

Method

Designation No.

Method Code"

Andersen Instruments FH62I-N Beta

EQPM-0990-076

076

DKK-TOA FPM-222/222C/223/223C

EQPM-0905-156

156

DKK-TOA DUB-222(S)/223(S)

EQPM-0905-156

156

Environnement S.A. MP101M Beta

EQPM-0404-151

151

Horiba APDA-371

EQPM-0798-122

122

Met One BAM1020, GBAM1020, BAM1020-1, GBAM1020-1 Beta

EQPM-0798-122

122

Met One Instruments, Inc. E-BAM PLUS

EQPM-1215-226

226

Opsis SM200

EQPM-0810-193

193

R & P TEOM® 1400, 1400a

EQPM-1090-079

079

SWAM 5a Dual Channel Monitor

EQPM-0912-205

205

Teledyne Model 602 BetaPLUS Particle Measurement System

EQPM-0912-205

205

Teledyne Model T640 PM Mass Monitor with 640X option

EQPM-0516-239

239

Thermo Andersen Series FH 62 C14 Beta Monitor, 5014i Beta

EQPM-1102-150

150

Thermo Scientific TEOM® 1400AB, 1405

EQPM-1090-079

079

Thermo Scientific TEOM® 1405-DF with FDMS®

EQPM-1013-208

208

W&A/Thermo Electron 650 Beta Gauge

EQPM-0391-081

081


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 55

Particulate Matter - PM2.5

PM2.5 Samplers

Method

Designation No/'

Method Code"

Andersen Model RAAS2.5-200 Audit with WINS

RFPS-0299-128

128

BGI or Mesa Laboratories Inc. PQ200/200A with WINS

RFPS-0498-116

116

BGI or Mesa Laboratories Inc. PQ200-VSCC™ or PQ200A-VSCC™

RFPS-0498-116 or
EQPM-0202-142

142

Graseby Andersen RAAS2.5-100 with WINS

RFPS-0598-119

119

Graseby Andersen RAAS2.5-300 with WINS

RFPS-0598-120

120

Met One Instruments, Inc. E-FRM with WINS

RFPS-0315-221

221

Met One Instruments, Inc E-FRM with VSCC™

RFPS-0315-221

521

R & P Partisol®-FRM 2000 PM2.5 with WINS

RFPS-0498-117

117

R & P Partisol®-FRM 2000 PM-2.5 Thermo Scientific with VSCC™

RFPS-0498-117 or
EQPM-0202-143

143

R & P Partisol® 2000 PM-2.5 Audit with WINS

RFPS-0499-129

129

R & P Partisol® 2000 PM-2.5 Audit with VSCC™

RFPS-0499-129 or
EQPM-0202-144

144

R & P Partisol®-Plus 2025 PM-2.5 Seq. with WINS

RFPS-0498-118

118

R & P Partisol®-Plus 2025 PM-2.5 Seq. with VSCC™

RFPS-0498-118 or
EQPM-0202-145

145

Thermo Electron RAAS2.5-100 with VSCC™

RFPS-0598-119 or
EQPM-0804-153

153

Thermo Electron RAAS2.5-100 with WINS

RFPS-0598-119

119

Thermo Electron RAAS2.5-200 with VSCC™

RFPS-0299-128 or
EQPM-0804-154

154

Thermo Electron RAAS2.5-200 with WINS

RFPS-0299-128

128

Thermo Electron RAAS2.5-300 with VSCC™

RFPS-0598-120 or
EQPM-0804-155

155

Thermo Electron RAAS2.5-300 with WINS

RFPS-0598-120

120

Thermo Environmental Model 605 CAPS

RFPS-1098-123

123

Thermo Scientific Partisol® 2000-D Dichot., Partisol® 2000/-D Dichot.

EQPS-0509-177

177

Thermo Scientific Dichot. Partisol®-Plus Model 2025-D Seq., Partisol®
2025/-D Dichot. Seq.

EQPS-0509-179

179

Thermo Scientific Partisol® 2000-FRM, Partisol® 2000/with WINS

RFPS-0498-117

117

Thermo Scientific Partisol® 2000-FRM, Partisol® 2000/with VSCC™

RFPS-0498-117 or
EQPM-0202-143

143

Thermo Scientific Partisol®-Plus 2025 Seq., Partisol® 2025/ Seq with
WINS

RFPS-0498-118

118

Thermo Scientific Partisol®-Plus 2025 Seq., Partisol® 2025/ Seq with
VSCC™

RFPS-0498-118 or
EQPM-0202-145

145

Tisch Environmental Model TE-Wilbur2.5 with WINS

RFPS-1014-219

219

Tisch Environmental Model TE-Wilbur 2.5 with VSCC™

RFPS-1014-219

519

Tisch Environmental Model TE-Wilbur 2.5 with Tisch TE-PM2.5C
cyclone

RFPS-1014-219 or
EQPS-0415-223

223

URG-MASS100 with WINS

RFPS-0400-135

135

URG-MASS100 with VSCC™

RFPS-0400-135

535

URG-MASS100 with URG-2000-30EGN cyclone

EQPS-0316-233

233


-------
PM2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 56

PM2.5 Samplers

Method

Designation No.''

Method Code"

URG-MASS300

RFPS-0400-136

136

URG-MASS300 with VSCC™

RFPS-0400-136

536

URG-MASS300 with URG-2000-30EGN cyclone

EQPS-0316-234

234

a For cases with two designation numbers, these methods are dual designated as FRM and FEM. However, where a method is dual
designated, EPA considers the method an FRM for purposes of ensuring QA requirements are met.
a Method code as available in the AQS data system



PM2.5 Analyzers

Method

Designation No.

Method Code"

Environnement s.a. Model MP101M PM2.5 Beta Attenuation Monitor

EQPM-1013-211

211

Grimm Model EDM 180 PM2.5 Monitor

EQPM-0311-195

195

Horiba APDA-371

EQPM-0308-170

170

Met One BAM-1020 PM-2.5 with VSCC™

EQPM-0308-170

170

Met One BAM-1020 PM-2.5 with TE-PM2.5C

EQPM-0308-170

570

Met One BAM-1022 Real Time Beta Attenuation Mass Monitor

EQPM-1013-209

209

Met One BAM-1022 with TE-PM2.5C

EQPM-1013-209

509

Met One BAM-1020 with URG-2000-30EGN Cyclone

EQPM-0715-266

266

Thermo Scientific FH62C14-DHS Continuous, 5014i

EQPM-0609-183

183

Opsis SM200

EQPM-0812-203

203

SWAM 5a Dual Channel Monitor

EQPM-0912-204

204

Teledyne Model 602 BetaPLUS Particle Measurement System

EQPM-0912-204

204

Teledyne Model T640 PM Mass Monitor

EQPM-0516-236

236

Teledyne Model T640 PM Mass Monitor with 640X option

EQPM-0516-238

238

Thermo Scientific Model 5030i Sharp, 5030 SHARP

EQPM-0609-184

184

Thermo Scientific TEOM® 1400a with Series 8500C FDMS®

EQPM-0609-181

181

Thermo Scientific TEOM® 1405-DF Dichot. with FDMS®

EQPM-0609-182

182

Tisch Environmental TE-EDM 180 PM2.5

EQPM-0311-195

195


-------
PMio-2.5

List of Designated Reference and Equivalent Methods. December 17.2016

Page 57

Particulate Matter - PM10-2.5

PM10-2.5 Samplers

Method

Designation No.

Method Code"

BGI or Mesa Laboratories, Inc. Model PQ200 PM10-2 5 sampler pair

RFPS-1208-173

173

Met One Instruments, Inc. E-FRM PM10-2.5

RFPS-0316-232

232

Thermo Scientific Partisol® 2000-FRM sampler pair, Partisol® 2000/
sampler pair

RFPS-0509-175

175

Thermo Scientific Partisol®-Plus 2025 Seq. sampler pair, Partisol® 2025/
Seq sampler pair

RFPS-0509-176

176

Thermo Scientific Partisol® 2000-D Dichot., Partisol® 2000/-D Dichot.

EQPS-0509-178

178

Thermo Scientific Dichot. Partisol®-Plus Model 2025-D Seq., Partisol®
2025/-D Dichot. Seq.

EQPS-0509-180

180

Tisch Environmental Model TE-Wilbur Low-Volume Air Particulate
Sampler Pair with WINS

RFPS-1014-220

220

Tisch Environmental Model TE-Wilbur Low-Volume Air Particulate
Sampler Pair with VSCC™

RFPS-1014-220

520

Tisch Environmental Model TE-Wilbur Low-Volume Air Particulate

RFPS-1014-220 or

224

Sampler Pair with Tisch TE-PM2.5C cyclone

EQPS-0415-224



PM10-2.5 Analyzers

Method

Designation No.

Method Code"

Met One Instruments BAM-1020 System

EQPM-0709-185

185

SWAM 5a Dual Channel Monitor

EQPM-0912-206

206

Teledyne Model 602 BetaPLUS Particle Measurement System

EQPM-0912-206

206

Teledyne Model T640 PM Mass Monitor with 640X option

EQPM-0516-240

240

Thermo Scientific TEOM® 1405-DF with FDMS®

EQPM-1013-207

207


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016

Page 58

Sulfur Dioxide

SO2 Manual Methods

Method

Designation No.

Method Code"

Reference method (pararosaniline)

—

097

Technicon 1 (pararosaniline)

EQS-0775-001

097

Technicon II (pararosaniline)

EQS-0775-002

097



SO2 Analyzers

Method

Designation No.

Method Code"

Advanced Pollution Instr. 100

EQSA-0990-077

077

Advanced Pollution Instr. 100A/100AS

EQSA-0495-100

100

Asarco 500

EQSA-0877-024

024

Beckman 953

EQSA-0678-029

029

Bendix 8303

EQSA-1078-030

030

Casella ML9850, ML9850B, CM2050, CM2050B

EQSA-0193-092

092

Columbia Scientific Industries 5700

EQSA-0494-095

095

Dasibi 4108

EQSA-1086-061

061

DKK-TOA Corp. GFS-32

EQSA-0701-115

115

DKK-TOA Corp. GFS-112E, GFS-112E-1

EQSA-0100-133

133

DKK-TOA Corp. GFS-312E

EQSA-1107-168

168

Ecotech ML9850/EC9850, ML9850B/EC9850B

EQSA-0193-092

092

Ecotech EC9850T

EQSA-0193-092

092

Ecotech Serinus 50

EQSA-0809-188

188

Environnement S.A AF21M

EQSA-0292-084

084

Environnement S.A AF22M

EQSA-0802-149

149

Environnement S.A. SANOA

EQSA-0400-138

138

Horiba Model APSA-360/APSA-360ACE

EQSA-0197-114

114

Horiba Model APSA-370

EQSA-0506-159

159

Lear Siegler AM2020

EQSA-1280-049

049

Lear Siegler SM1000

EQSA-1275-005

005

Meloy SA185-2A

EQSA-1275-006

006

Meloy SA285E

EQSA-1078-032

032

Meloy SA700

EQSA-0580-046

046

Monitor Labs 8450

EQSA-0876-013

513

Monitor Labs or Lear Siegler 8850

EQSA-0779-039

039

Monitor Labs or Lear Siegler 8850S

EQSA-0390-075

075

Opsis AB OPS50

EQSA-0809-188

188

Opsis AR 500, System 300 (open path)

EQSA-0495-101

101

Philips PW9700

EQSA-0876-011

511

Philips PW9755

EQSA-0676-010

010

recordum airpointer® system

EQSA-0495-100

100

recordum airpointer® system

EQSA-0486-060

060

SERES SF 2000 G

EQSA-0810-194

194

SIR S.A. S-5001

EQSA-0507-166

166

Sutron Model 6020

RFSA-0616-237

237

Teledyne Advanced Pollution Inst. 100A, 100 AS, 100E, 100EU, T100,
T100U

EQSA-0495-100

100


-------
so2

List of Designated Reference and Equivalent Methods. December 17.2016

Page 59

SO2 Analyzers

Method

Designation No.

Method Code"

Teledyne Analytical Instruments 6400A

EQSA-0495-100

100

Teledyne Analytical Instruments 6400E

EQSA-0809-188

188

Teledyne Monitor Labs ML9850, ML9850B

EQSA-0193-092

092

Teledyne Monitor Labs TML-50

EQSA-0495-100

100

Thermo Electron 43

EQSA-0276-009

009

Thermo Electron 43A, 43C-TLE, 43/

EQSA-0486-060

060

Thermo Environmental Instruments 43B, 43C

EQSA-0486-060

060

Tisch Environmental Model TE 2.0

EQSA-0809-188

188

Wedding 1040

EQSA-0193-092

092


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 60

Ozone

O3 Analyzers

Method

Designation No.

Method Code15

2B Technologies Model 202

EQOA-0410-190

190

2B Technologies Model 211

EQOA-0514-215

215

2B Technologies Model 106

EQOA-0914-218

218

2B Technologies Model POM

EQOA-0815-227

227

Advanced Pollution Instr. 400/400A/400E

EQOA-0992-087

087

Beckman 950A

RFOA-0577-020

020

Bendix 8002

RFOA-0176-007

007

Casella ML9810, ML9810B, ML9811, ML9812, CM2010, CM2010B,
CM2011, CM2012

EQOA-0193-091

091

Columbia Scientific Industries 2000

RFOA-0279-036

036

Dasibi 1003-AH, -PC, -RS

EQOA-0577-019

019

Dasibi 1008-AH, -PC, -RS

EQOA-0383-056

056

DKK-TOA Corp. GUX-113E, GUX-113E-1

EQOA-0200-134

134

DKK-TOA Corp. GUX-313E

EQOA-1107-169

169

Ecotech ML9810/EC9810, -9810B, -9811, -9812

EQOA-0193-091

091

Ecotech Serinus 10

EQOA-0809-187

187

Environics 300

EQOA-0990-078

078

Environnement S.A 0341M

EQOA-0895-105

105

Environnement S.A 0342M

EQOA-0206-148

148

Environnement S.A SANOA

EQOA-0400-137

137

Environnement S.A O3 42e

EQOA-0515-225

225

Horiba APOA-360

EQOA-0196-112

112

Horiba APOA-370

EQOA-0506-160

160

McMillan 1100-1

RFOA-1076-014

514

McMillan 1100-2

RFOA-1076-015

515

McMillan 1100-3

RFOA-1076-016

016

Meloy OA325-2R

RFOA-1075-003

003

Meloy OA350-2R

RFOA-1075-004

004

Monitor Labs 8410E

RFOA-1176-017

017

Monitor Labs or Lear Siegler 8810

EQOA-O881-053

053

Opsis AB OPS10

EQOA-0809-187

187

Opsis AR 500, System 300 (open path)

EQOA-0495-103

103

PCI Ozone Corp. LC-12

EQOA-0382-055

055

Philips PW9771

EQOA-0777-023

023

recordum airpointer® system

EQOA-0992-087

087

recordum airpointer® system

EQOA-0880-047

047

Seres Model OZ 2000 G

EQOA-0506-161

161

SIR S.A. S-5014

EQOA-0207-164

164

Sutron Model 6030

EQOA-0415-222

222

Tanabyte 722, 723, 724, 725, 726

EQOA-0407-165

165

Teledyne Advanced Pollution Instr. 265E, T265

EQOA-0611-199 or
RFOA-0216-230

199

Teledyne Advanced Pollution Instr. 400, 400A, 400E, T400

EQOA-0992-087

087

Teledyne Advanced Pollution Instr. T204

EQOA-0514-214

214

Teledyne Advanced Pollution Instr. Model 430

EQOA-1015-229

229

Teledyne Monitor Labs ML9810/9810B, ML9811, ML9812

EQOA-0193-091

091

Teledyne Monitor Labs TML-10

EQOA-0992-087

087


-------
o3

List of Designated Reference and Equivalent Methods. December 17.2016

Page 61

O3 Analyzers

Method

Designation No.

Method Code"

Thermo Electron or Thermo Environmental Instruments 49, 49C, 49/

EQOA-0880-047

047

Tisch Environmental TE 1.0

EQOA-0809-187

187

Wedding 1010

EQOA-0193-091

091


-------
CO

List of Designated Reference and Equivalent Methods. December 17.2016

Page 62

Carbon Monoxide

CO Analyzers

Method

Designation No.

Method Code"

Beckman 866

RFCA-0876-012

012

Bendix 8501-5CA

RFCA-0276-008

008

Casella ML9830, ML9830B, CM2030, CM2030B

RFCA-0992-088

088

Dasibi 3003

RFCA-0381-051

051

Dasibi 3008

RFCA-0488-067

067

DKK-TOA Corp. GFC-311E

RFCA-0907-167

167

Ecotech Serinus 30

RFCA-0509-174

174

Ecotech ML9830/EC9830, ML9830B/EC9830B

RFCA-0992-088

088

Ecotech EC9830T

RFCA-0992-088

088

Environnement S.A CO11M

RFCA-0995-108

108

Environnement S.A C012M

RFCA-0206-147

147

Environnement S.A C012e

RFCA-0915-228

228

Horiba AQM-10, -11, -12

RFCA-1278-033

033

Horiba 300E/300SE

RFCA-1180-048

048

Horiba APMA-360

RFCA-0895-106

106

Horiba APMA-370

RFCA-0506-158

158

MASS - CO 1 (Massachusetts)

RFCA-1280-050

050

Monitor Labs 8310

RFCA-0979-041

041

Monitor Labs or Lear Siegler 8830

RFCA-0388-066

066

MSA 202S

RFCA-0177-018

018

Opsis AB OPS 30

RFCA-0509-174

174

Peak Laboratories Model 910-170

EQCA-0814-217

217

recordum airpointer® system

RFCA-1093-093

093

recordum airpointer® system

RFCA-0981-054

054

SIR S.A. Model S-5006

RFCA-0708-172

172

Teledyne Adv. Pollution Instr. 300, 300E, 300EU, T300, T300U

RFCA-1093-093

093

Teledyne Analytical Instruments GFC7001E

RFCA-0509-174

174

Teledyne Monitor Labs ML9830/9830B

RFCA-0992-088

088

Teledyne Monitor Labs TML-30

RFCA-1093-093

093

Thermo Electron or Thermo Environmental Instruments 48, 48C, 48/',
48/TLE

RFCA-0981-054

054

Tisch Environmental Model TE 3.0

RFCA-0509-174

174

Wedding 1020

RFCA-0992-088

088


-------
List of Designated Reference and Equivalent Methods. December 17.2016

Page 63

Nitrogen Dioxide

NO2 Manual Methods

Method

Designation No.

Method Code"

Sodium arsenite (orifice)

EQN-1277-026

084

Sodium arsenite/Technicon II

EQN-1277-027

084

TGS-ANSA (orifice)

EQN-1277-028

098



NO2 Analyzers

Method

Designation No.

Method Code"

Advanced Pollution Instr. 200

RFNA-0691-082

082

Advanced Pollution Instr. 200A/200AU

RFNA-1194-099

099

Beckman 952A

RFNA-0179-034

034

Bendix 8101-B

RFNA-0479-038

038

Bendix 8101-C

RFNA-0777-022

022

Casella ML9841, ML9841A, ML9841B, CM2041, CM2041A, CM2041B

RFNA-1292-090

090

Columbia Scientific lndust.1600, 5600

RFNA-0977-025

025

Dasibi 2108

RFNA-1192-089

089

DKK-TOA Corp GLN-114E, GLN-114E-1

RFNA-0798-121

121

DKK-TOA Corp GLN-314E

RFNA-0508-171

171

Ecotech ML9841A/EC9841 A.ML9841 B/EC9841B

RFNA-1292-090

090

Ecotech EC9841T

RFNA-1292-090

090

Ecotech Serinus 40

RFNA-0809-186

186

Environnement S.A. AC31M

RFNA-0795-104

104

Environnement S.A. AC32M

RFNA-0202-146

146

Environnement S.A. SANOA

EQNA-0400-139

139

Environnement S.A. AS32M

EQNA-1013-210

210

Horiba APNA-360

RFNA-0196-111

111

Horiba APNA-370

RFNA-0506-157

157

Meloy NA530R

RFNA-1078-031

031

Monitor Labs 8440E

RFNA-0677-021

021

Monitor Labs or Lear Siegler 8840

RFNA-0280-042

042

Monitor Labs or Lear Siegler 8841

RFNA-0991-083

083

Monitor Labs ML9841

RFNA-1292-090

090

Opsis AB OPS40

RFNA-0809-186

186

Opsis AR 500, System 300 (open path)

EQNA-0495-102

102

Philips PW9762/02

RFNA-0879-040

040

recordum airpointer system

RFNA-1194-099

099

recordum airpointer system

RFNA-1289-074

074

Seres Model NOx 2000 G

RFNA-0706-163

163

SIR S.A. S-5012

RFNA-0804-152

152

Teledyne Advanced Pollution Inst. 200A, 200AU, 200E, 200EU, T200,
T200U, T204

RFNA-1194-099

099

Teledyne Advanced Pollution Inst. T500U

EQNA-0514-212

212

Teledyne Advanced Pollution Inst. 200EUP, T200UP

EQNA-0512-200

200

Teledyne Advanced Pollution Inst. T200P

EQNA-1016-241

241

Teledyne Analytical Instruments 911 OA

RFNA-1194-099

099

Teledyne Analytical Instruments 9110E

RFNA-0809-186

186


-------
List of Designated Reference and Equivalent Methods. December 17.2016

Page 64

NO2 Analyzers

Method

Designation No.

Method Code"

Teledyne Monitor Labs ML9841, ML9841A, ML9841B

RFNA-1292-090

090

Teledyne Monitor Labs TML-41

RFNA-1194-099

099

Thermo Electron or Thermo Environmental Instruments 14B/E

RFNA-0179-035

035

Thermo Electron or Thermo Environmental Instruments 14D/E

RFNA-0279-037

037

Thermo Environmental Instr. 42, 42C, 42/

RFNA-1289-074

074

Tisch Environmental Model TE 4.0

RFNA-0809-186

186

Wedding 1030

RFNA-1292-090

090


-------
Pb

List of Designated Reference and Equivalent Methods. December 17.2016

Page 65

Lead

Pb Manual Methods

Method

Designation No.

Method Code"

H

-vol/AAS (alt. extr.)

EQL-0380-043

043

H

-vol/Energy-disp XRF (TX ACB)

EQL-0783-058

058

H

-vol/Energy-disp XRF (NEA)

EQL-0589-072

072

H

-vol/FAAS

EQLA-0813-803

803

H

-vol/Flameless AA (EMSL/EPA)

EQL-0380-044

044

H

-vol/Flameless AA (Houston)

EQL-0895-107

107

H

-vol/Flameless AA (Omaha)

EQL-0785-059

059

H

-vol/ICAP-OES (Doe Run Co.)

EQL-0196-113

113

H

-vol/ICAP-OES (EMSL/EPA)

EQL-0380-045

045

H

-vol/ICAP-OES (Illinois)

EQL-1193-094

094

H

-vol/ICAP-OES (Kansas)

EQL-0592-085

085

H

-vol/ICAP-OES (Montana)

EQL-0483-057

057

H

-vol/ICAP-OES (NE&T)

EQL-1188-069

069

H

-vol/ICAP-OES (New Hampshire)

EQL-1290-080

080

H

-vol/ICAP-OES (Pennsylvania)

EQL-0592-086

086

H

-vol/ICAP-OES (Pima Co.,AZ)

EQL-0995-109

109

H

-vol/ICAP-MS (Pima Co.,AZ)

EQL-0995-110

110

H

-vol/ICAP-OES (Rhode Island)

EQL-0888-068

068

H

-vol/ICAP-OES (Silver Val. Labs)

EQL-1288-070

070

H

-vol/ICAP-AES (TNRCC)

EQL-0400-140

140

H

-vol/ICAP-OES (West Virginia)

EQL-0694-096

096

H

-vol/WL-disp. XRF (CA A&IHL)

EQL-0581-052

052

H

-vol/ICP-MS (IML, Inc)

EQL-0310-189

189

H

-Vol/ICP-MS (US EPA/OAQPS)

EQL-0510-191

191

H

-Vol/ICP-MS (US EPA/Region 9)

EQL-0710-192

192

H

-Vol/ICP-AES (US EPA/OAQPS)

EQL-0311-196

196

H

-Vol/ICP-MS (ERG)

EQL-0512-201

201

PM10/ICP-MS (ERG)

EQL-0512-202

202

Hi-Vol/TOF ICP-MS (SCAQMD)

EQL-0514-213

213


-------
Sources/Contacts List of Designated Reference and Equivalent Methods. December 17.2016 Page 66

Sources or Contacts for Designated Reference and Equivalent Methods

2B Technologies, Inc.

2100 Central Ave., Suite 105
Boulder, CO 80301

(303)273-0559
http://www.twobtech.com

ABB Process Analytics
P.O. Box 831
Lewisburg, WV 24901

(304)	647-4358

Advanced Pollution Instrumentation, Inc.

[Refer to Teledyne Advanced Pollution
Instrumentation]

Andersen Instruments

[Refer to Thermo Fisher Scientific, Inc.]

ASARCO Incorporated

3422 South 700 West
Salt Lake City, UT 84119
(801)262-2459

Beckman Instruments, Inc.

Process Instruments Division
2500 Harbor Blvd.

Fullerton, CA 92634
(714)871-4848

Bendix

[Refer to ABB Process Analytics]

BGI Incorporated

(Refer to Mesa Labs, Inc.)

Casella Monitor

Regent House, Wolseley Road, Kempston
Bedford, United Kingdom MK42 7JY
www.casellameasurement.com

Columbia Scientific Industries

11950 Jollyville Road
Austin, TX 78759
(800) 531-5003

Combustion Engineering

[Refer to ABB Process Analytics]

Dasibi Environmental Corp.

[Formerly, 506 Paula Avenue
Glendale, CA 91201
(818) 247-7601
www.dasibi.com

DKK-TOA Corporation

29-10, 1-Chome, Takadanobaba,

Shinjuku-ku

Tokyo 169-8648, Japan

+81-3-3202-0211

www.toadkk.co.ip

Eastern Research Group, Inc.

601 Keystone Park Drive, Suite 700,
Morrisville, NC 27560
919- 468-7800
www.erg.com

Ecotech Pty. Ltd.

1492 Ferntree Gully Rd.

Knoxfield, Victoria, 3180, Australia
+61 1300 364 946
www.ecotech.com.au

Environnement S.A.

111, bd Robespierre
78300 Poissy, France
www.environnement-sa.com
Instruments also available from:
Altech/Environnement U.S.A.
2623 Kaneville Court
Geneva, IL 60134
(630) 262- 4400

Environics, Inc.

69 Industrial Park Rd. E.

Tolland, CT 06084-2805
(203) 429-0077
www.environics.com

FAI Instruments S.R.L

Via Aurora, 15 -00013 FonteNuova
Rome, Italy
(+39) 06.9050248
www.fai-instruments.it

Graseby GMW

[Refer to Thermo Fisher Scientific, Inc.]

GRIMM Technologies, Inc.

5833 Stewart Parkway, Suite 203
Douglasville, GA 30153
(877) 474-6872
www.dustmonitor.com

Horiba Instruments Incorporated

17671 Armstrong Avenue
Irvine, CA 92714
(800)446-7422
www.horiba.com

Inter-Mountain Labs, Inc.

1673 Terra Avenue
Sheridan, WY 82801
(307) 672-8945
www.intermountainlabs.com

Lear Siegler

[Refer to Teledyne Monitor Labs, Inc.]

Commonwealth of Massachusetts

Department of Environmental
Quality Engineering
Tewksbury, MA 01876

Mesa Laboratories Inc.

10 Park Place
Butler, NC 07405
(973) 492-8400
www.mesalabs.com

Met One Instruments, Inc.

1600 Washington Blvd.

Grants Pass, OR 97526
(541)471-7111

www.metone.com(metone@metone.com)
McMillan

[Refer to Columbia Scientific Industries]

Mine Safety Appliances

600 Penn Center Blvd.

Pittsburgh, PA 15235-5810
(412) 273-5101

Monitor Labs, Inc.

[Refer to Teledyne Monitor Labs, Inc.]


-------
Sources/Contacts List of Designated Reference and Equivalent Methods. December 17.2016 Page 67

New Star Environmental, LLC

3293 Ashburton Chase NE
Roswell, GA 30075
(770)998-2590

Opsls AB

Box 244

SE-244 02 Furalund
Sweden

+46 46 72 25 00
www.opsis.se

Instruments also available from
Opsls, Inc.

150 N. Michigan Ave., Suite 1950
Chicago, IL 60601
(312) 447-7733

State of Oregon

Department of Environmental Quality
Air Quality Division
811 S.W. Sixth Avenue
Portland, OR 97204

PCI Ozone Corp.

One Fairfield Crescent
West Caldwell, NJ 07006
(201)575-7052
www.pci-wedeco.com

Peak Laboratories, LLC

2330 Old Middlefield Way Suite 10
Mountain View, CA 94043
650-691-1267
www.peaklaboratories.com

Phillips Electronic Instruments, Inc.

85 McKee Drive
Mahwall, NJ 07430

Recordum Messtechnik GmbH

Haus 1/Top 403
Triester Str. 14
A-2351 Wiener Neudorf
Austria

www.recordum.com

Rupprecht & PatashnickCo., Inc.

[Refer to Thermo Fisher Scientific, Inc.]

Slbata Scientific Technology, Ltd.

1-25, 3-chome
Ikenohata, Taito-ku
Tokyo 110, Japan
81-3(3822)2272
TTani@email.msn.com

Seres

360 Rue Louis de Broglie
La Duranne - BP 87000
13793 Aix en Provence - Cedex 3
France

+33 (0)4 42 97 37 37
www.seres-france.com

SIR SlA. (Sistemas Iratalaaones y Redes, SA.)

Avenida de la Industria, 3
Tres Cantos 28760 Madrid, SPAIN
(+34) 91 803 66 02
www.sirsa.es

SIR USA

826 West Braddock Road
Alexandria, VA 22302-3605
(703)837-1883

South Coast Air Quality Management
District

21865 Copley Drive
Diamond Bar, CA 91765-4182
(909) 396-2269
ilow@aqmd.gov

Sutron Air Quality Division

2548 Shell Road
Georgetown, TX 78628.

512-869-0544
www, sutron. com/aq

Tanabyte Engineering, Inc.

1210 West Burbank Blvd., Suite B,
Burbank, CA 91506
(818) 842-4022
www.tanabvte.com

Teledyne Advanced Pollution
Instrumentation

9970 Carroll Canyon Rd, Suite A
San Diego, CA 92131
(858)657-9800
www.teledvne-api .com

Teledyne Analytical Instruments

16830 Chestnut Street
City of Industry, CA 91748
(626) 934-1622

Teledyne Monitor Labs, Inc.

74 Inverness Drive East
Englewood, CO 80112-5189
(303) 792-3300
www.teledvne-ml.com

Thermo Electron Corporation

[Refer to Thermo Fisher Scientific, Inc.]

Thermo Environmental Instruments, Inc.

[Refer to Thermo Fisher Scientific, Inc.]

Thermo Fisher Scientific, Inc.

81 Wyman Street

Waltham, MA 02454

(781) 622-1000 / (800) 678-5599

www.thermo.com

Tisch Environmental, Inc.

145 S. Miami Avenue
Village of Cleves, OH 45002
(513)467-9000
www.tisch-env.com

URG Corporation

116 Merritt Mill Road
Chapel Hill, NC 27516
(919) 942-2753

U.S. EPA

National Exposure Research Laboratory
Human Exposure & Atmospheric
Sciences Division
Process Modeling Research Branch
(MD D205-03)

Research Triangle Park, NC 27711

(919)541-5691

www.epa.gov/heasd

U.S. EPA

Office of Air and Radiation
Air Quality Planning and Standards
Air Quality Assessment Division
(MD C404-03)

Research Triangle Park, NC 27711

(919)541-3372

www.epa.gov/air

U.S. EPA

Region 9 Laboratory

1337 South 46th Street, Bldg 201

Richmond, CA 94804

Wedding and Associates, Inc.

[Refer to Thermo Fisher Scientific, Inc.]


-------